SLAC Magnetic Measurements Date: 03-15-2002 Time: 12:25:51 Project: spear3 Magnet Type: quad Magnet Name: 50q-095 Serial Number: Run Number: 2 Measurement Device: ten-turn wire Operator: zw Comment: GL vs I for transfer function Standardization: 3 cycles from 0.0 A to 98.0 A to 0.0 A Test Currents (A): 5.0 10.0 15.0 20.0 25.0 30.0 35.0 40.0 45.0 50.0 55.0 60.0 65.0 70.0 75.0 80.0 85.0 90.0 80.0 70.0 60.0 50.0 40.0 30.0 20.0 10.0 12:26:27 Initial Zero Current Measurement, Imag_zero = 0.0000 A 12:26:27 Magnet current ramping to 5.0000 A... 12:26:55 Magnet current ramping to 98.0000 A... 12:27:28 Magnet current ramping to 0.0000 A... 12:28:02 Magnet current ramping to 98.0000 A... 12:28:37 Magnet current ramping to 0.0000 A... 12:29:11 Magnet current ramping to 98.0000 A... 12:29:46 Magnet current ramping to 0.0000 A... 12:30:21 Magnet current ramping to 5.0000 A... 12:30:31 Magnet Current, Imag = 4.9060 A Vtrans = 0.049060 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 12:30:33 Magnet Current, Imag = 4.9193 A Vtrans = 0.049193 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 12:30:35 Magnet Current, Imag = 4.9169 A Vtrans = 0.049169 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 12:30:36 Magnet Current, Imag = 4.9167 A Vtrans = 0.049167 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 12:30:36 Magnet Current Measurement: Imag = 4.9147 +- 0.0052 A 12:30:36 Magnetic Field Measurement Measuring the magnet strength... 12:30:36 Integrated Voltage Measurement... 12:30:59 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = 0.0100000 m VT = -0.001293117 Vs 12:31:14 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = -0.0100000 m VT = 0.001398414 Vs 12:31:29 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = 0.0100000 m VT = -0.001284646 Vs 12:31:44 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = -0.0100000 m VT = 0.001414152 Vs 12:31:59 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = 0.0100000 m VT = -0.001273380 Vs 12:32:14 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = -0.0100000 m VT = 0.001421458 Vs 12:32:29 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = 0.0100000 m VT = -0.001273390 Vs 12:32:44 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = -0.0100000 m VT = 0.001418781 Vs 12:32:44 Integrated Voltage Average and RMS of Measurements X0 = -0.0200000 m, Delta X = 0.0100000 m VT = -0.001347167 +- 0.000001430 Vs 12:32:44 Integrated Field Strength, Average and RMS of Measurements X0 = -0.0200000 m, dx = 0.0100000 m, Nturns = 10.0 VT = -0.001347167 +- 0.000001430 Vs BL = -0.01347167 +- 0.00001430 Tm 12:32:44 Integrated Voltage Measurement... 12:33:03 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = 0.0100000 m VT = -0.000616544 Vs 12:33:18 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = -0.0100000 m VT = 0.000770859 Vs 12:33:33 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = 0.0100000 m VT = -0.000610298 Vs 12:33:48 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = -0.0100000 m VT = 0.000771628 Vs 12:34:03 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = 0.0100000 m VT = -0.000615440 Vs 12:34:18 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = -0.0100000 m VT = 0.000769198 Vs 12:34:33 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = 0.0100000 m VT = -0.000624311 Vs 12:34:48 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = -0.0100000 m VT = 0.000762350 Vs 12:34:48 Integrated Voltage Average and RMS of Measurements X0 = -0.0100000 m, Delta X = 0.0100000 m VT = -0.000692578 +- 0.000001061 Vs 12:34:48 Integrated Field Strength, Average and RMS of Measurements X0 = -0.0100000 m, dx = 0.0100000 m, Nturns = 10.0 VT = -0.000692578 +- 0.000001061 Vs BL = -0.00692578 +- 0.00001061 Tm 12:34:48 Integrated Voltage Measurement... 12:35:06 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.000024446 Vs 12:35:22 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = 0.000100738 Vs 12:35:37 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.000016473 Vs 12:35:52 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = 0.000100929 Vs 12:36:07 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.000021912 Vs 12:36:22 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = 0.000102482 Vs 12:36:37 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.000024152 Vs 12:36:52 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = 0.000108254 Vs 12:36:52 Integrated Voltage Average and RMS of Measurements X0 = 0.0000000 m, Delta X = 0.0100000 m VT = -0.000040677 +- 0.000001647 Vs 12:36:52 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0000000 m, dx = 0.0100000 m, Nturns = 10.0 VT = -0.000040677 +- 0.000001647 Vs BL = -0.00040677 +- 0.00001647 Tm 12:36:52 Integrated Voltage Measurement... 12:37:11 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = 0.0100000 m VT = 0.000673823 Vs 12:37:26 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = -0.0100000 m VT = -0.000546076 Vs 12:37:41 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = 0.0100000 m VT = 0.000663696 Vs 12:37:56 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = -0.0100000 m VT = -0.000560030 Vs 12:38:11 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = 0.0100000 m VT = 0.000658021 Vs 12:38:26 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = -0.0100000 m VT = -0.000553522 Vs 12:38:41 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = 0.0100000 m VT = 0.000666877 Vs 12:38:57 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = -0.0100000 m VT = -0.000548559 Vs 12:38:57 Integrated Voltage Average and RMS of Measurements X0 = 0.0100000 m, Delta X = 0.0100000 m VT = 0.000608825 +- 0.000002294 Vs 12:38:57 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0100000 m, dx = 0.0100000 m, Nturns = 10.0 VT = 0.000608825 +- 0.000002294 Vs BL = 0.00608825 +- 0.00002294 Tm 12:38:57 Integrated Voltage Measurement... 12:39:15 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = 0.0100000 m VT = 0.001312644 Vs 12:39:30 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = -0.0100000 m VT = -0.001200791 Vs 12:39:46 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = 0.0100000 m VT = 0.001308132 Vs 12:40:01 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = -0.0100000 m VT = -0.001209124 Vs 12:40:16 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = 0.0100000 m VT = 0.001293248 Vs 12:40:31 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = -0.0100000 m VT = -0.001224039 Vs 12:40:46 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = 0.0100000 m VT = 0.001284670 Vs 12:41:01 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = -0.0100000 m VT = -0.001228797 Vs 12:41:01 Integrated Voltage Average and RMS of Measurements X0 = 0.0200000 m, Delta X = 0.0100000 m VT = 0.001257681 +- 0.000000955 Vs 12:41:01 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0200000 m, dx = 0.0100000 m, Nturns = 10.0 VT = 0.001257681 +- 0.000000955 Vs BL = 0.01257681 +- 0.00000955 Tm 12:41:06 BL vs X Fit Coefficients: Cn sigCn n (Tm/m^n) ----- --------+-------- 0 -0.00043 0.00001 1 0.65111 0.00043 Compute Fit Residuals: X0 BL sigBL BLfit BL-BLfit (m) (Tm) (Tm) (Tm) (Tm) ---------- --------+-------- -------- -------- -0.020000 -0.01347 0.00001 -0.01345 -0.00002 -0.010000 -0.00693 0.00001 -0.00694 0.00001 0.000000 -0.00041 0.00002 -0.00043 0.00002 0.010000 0.00609 0.00002 0.00608 0.00000 0.020000 0.01258 0.00001 0.01259 -0.00002 12:41:06 Magnet Strength Calculation Harmonic Number = 2 Number Of X0 Positions = 5 X0 BL sigBL (m) (Tm) (Tm) -------- --------+-------- -0.0200 -0.01347 0.00001 -0.0100 -0.00693 0.00001 0.0000 -0.00041 0.00002 0.0100 0.00609 0.00002 0.0200 0.01258 0.00001 SL = 0.6511100 +- 0.0004268 Tm^2/m^n 12:41:06 Quadrupole Strength Calculation Check Slope = 0.651110 Tm/m, Inter = -0.000428 Tm, Sig_slope = 0.000687 Tm/m, Est_sig = 0.000022 Tm GL = 0.651110 +- 0.000687 T 12:41:06 Magnet Strength Measurement Harmonic Number = 2 SL = 0.6511100 +- 0.0004268 Tm^2/m^n 12:41:08 Magnet Current, Imag = 4.9165 A Vtrans = 0.049165 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 12:41:10 Magnet Current, Imag = 4.9166 A Vtrans = 0.049166 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 12:41:12 Magnet Current, Imag = 4.9160 A Vtrans = 0.049160 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 12:41:14 Magnet Current, Imag = 4.9171 A Vtrans = 0.049171 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 12:41:14 Magnet Current Measurement: Imag = 4.9165 +- 0.0004 A 12:41:14 Magnet current ramping to 10.0000 A... 12:41:23 Magnet Current, Imag = 10.0143 A Vtrans = 0.100143 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 12:41:25 Magnet Current, Imag = 10.0105 A Vtrans = 0.100105 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 12:41:27 Magnet Current, Imag = 10.0103 A Vtrans = 0.100103 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 12:41:29 Magnet Current, Imag = 10.0109 A Vtrans = 0.100110 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 12:41:29 Magnet Current Measurement: Imag = 10.0115 +- 0.0016 A 12:41:29 Magnetic Field Measurement Measuring the magnet strength... 12:41:29 Integrated Voltage Measurement... 12:41:51 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = 0.0100000 m VT = -0.002674160 Vs 12:42:06 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = -0.0100000 m VT = 0.002725387 Vs 12:42:21 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = 0.0100000 m VT = -0.002660266 Vs 12:42:36 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = -0.0100000 m VT = 0.002731385 Vs 12:42:51 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = 0.0100000 m VT = -0.002659728 Vs 12:43:06 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = -0.0100000 m VT = 0.002741014 Vs 12:43:21 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = 0.0100000 m VT = -0.002652845 Vs 12:43:36 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = -0.0100000 m VT = 0.002741163 Vs 12:43:36 Integrated Voltage Average and RMS of Measurements X0 = -0.0200000 m, Delta X = 0.0100000 m VT = -0.002698243 +- 0.000001888 Vs 12:43:36 Integrated Field Strength, Average and RMS of Measurements X0 = -0.0200000 m, dx = 0.0100000 m, Nturns = 10.0 VT = -0.002698243 +- 0.000001888 Vs BL = -0.02698243 +- 0.00001888 Tm 12:43:36 Integrated Voltage Measurement... 12:43:55 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = 0.0100000 m VT = -0.001346856 Vs 12:44:10 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = -0.0100000 m VT = 0.001419463 Vs 12:44:25 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = 0.0100000 m VT = -0.001350461 Vs 12:44:40 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = -0.0100000 m VT = 0.001417153 Vs 12:44:55 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = 0.0100000 m VT = -0.001355345 Vs 12:45:10 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = -0.0100000 m VT = 0.001417447 Vs 12:45:25 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = 0.0100000 m VT = -0.001356525 Vs 12:45:40 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = -0.0100000 m VT = 0.001423224 Vs 12:45:40 Integrated Voltage Average and RMS of Measurements X0 = -0.0100000 m, Delta X = 0.0100000 m VT = -0.001385809 +- 0.000002641 Vs 12:45:40 Integrated Field Strength, Average and RMS of Measurements X0 = -0.0100000 m, dx = 0.0100000 m, Nturns = 10.0 VT = -0.001385809 +- 0.000002641 Vs BL = -0.01385809 +- 0.00002641 Tm 12:45:40 Integrated Voltage Measurement... 12:45:59 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = -0.000042595 Vs 12:46:14 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = 0.000112485 Vs 12:46:29 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = -0.000043897 Vs 12:46:44 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = 0.000119298 Vs 12:47:00 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = -0.000039866 Vs 12:47:15 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = 0.000109374 Vs 12:47:30 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = -0.000054083 Vs 12:47:45 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = 0.000104279 Vs 12:47:45 Integrated Voltage Average and RMS of Measurements X0 = 0.0000000 m, Delta X = 0.0100000 m VT = -0.000078234 +- 0.000002537 Vs 12:47:45 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0000000 m, dx = 0.0100000 m, Nturns = 10.0 VT = -0.000078234 +- 0.000002537 Vs BL = -0.00078234 +- 0.00002537 Tm 12:47:45 Integrated Voltage Measurement... 12:48:03 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = 0.0100000 m VT = 0.001260836 Vs 12:48:19 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = -0.0100000 m VT = -0.001196149 Vs 12:48:34 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = 0.0100000 m VT = 0.001250615 Vs 12:48:49 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = -0.0100000 m VT = -0.001201675 Vs 12:49:04 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = 0.0100000 m VT = 0.001254179 Vs 12:49:19 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = -0.0100000 m VT = -0.001191684 Vs 12:49:34 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = 0.0100000 m VT = 0.001262165 Vs 12:49:49 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = -0.0100000 m VT = -0.001186398 Vs 12:49:49 Integrated Voltage Average and RMS of Measurements X0 = 0.0100000 m, Delta X = 0.0100000 m VT = 0.001225463 +- 0.000002089 Vs 12:49:49 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0100000 m, dx = 0.0100000 m, Nturns = 10.0 VT = 0.001225463 +- 0.000002089 Vs BL = 0.01225463 +- 0.00002089 Tm 12:49:49 Integrated Voltage Measurement... 12:50:07 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = 0.0100000 m VT = 0.002576190 Vs 12:50:22 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = -0.0100000 m VT = -0.002491257 Vs 12:50:38 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = 0.0100000 m VT = 0.002570107 Vs 12:50:53 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = -0.0100000 m VT = -0.002493491 Vs 12:51:08 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = 0.0100000 m VT = 0.002584593 Vs 12:51:23 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = -0.0100000 m VT = -0.002472700 Vs 12:51:38 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = 0.0100000 m VT = 0.002594500 Vs 12:51:53 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = -0.0100000 m VT = -0.002466363 Vs 12:51:53 Integrated Voltage Average and RMS of Measurements X0 = 0.0200000 m, Delta X = 0.0100000 m VT = 0.002531150 +- 0.000001859 Vs 12:51:53 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0200000 m, dx = 0.0100000 m, Nturns = 10.0 VT = 0.002531150 +- 0.000001859 Vs BL = 0.02531150 +- 0.00001859 Tm 12:51:58 BL vs X Fit Coefficients: Cn sigCn n (Tm/m^n) ----- --------+-------- 0 -0.00081 0.00001 1 1.30701 0.00063 Compute Fit Residuals: X0 BL sigBL BLfit BL-BLfit (m) (Tm) (Tm) (Tm) (Tm) ---------- --------+-------- -------- -------- -0.020000 -0.02698 0.00002 -0.02695 -0.00003 -0.010000 -0.01386 0.00003 -0.01388 0.00002 0.000000 -0.00078 0.00003 -0.00081 0.00003 0.010000 0.01225 0.00002 0.01226 -0.00000 0.020000 0.02531 0.00002 0.02533 -0.00002 12:51:58 Magnet Strength Calculation Harmonic Number = 2 Number Of X0 Positions = 5 X0 BL sigBL (m) (Tm) (Tm) -------- --------+-------- -0.0200 -0.02698 0.00002 -0.0100 -0.01386 0.00003 0.0000 -0.00078 0.00003 0.0100 0.01225 0.00002 0.0200 0.02531 0.00002 SL = 1.3070059 +- 0.0006278 Tm^2/m^n 12:51:58 Quadrupole Strength Calculation Check Slope = 1.307006 Tm/m, Inter = -0.000811 Tm, Sig_slope = 0.000941 Tm/m, Est_sig = 0.000030 Tm GL = 1.307006 +- 0.000941 T 12:51:58 Magnet Strength Measurement Harmonic Number = 2 SL = 1.3070059 +- 0.0006278 Tm^2/m^n 12:52:00 Magnet Current, Imag = 10.0115 A Vtrans = 0.100115 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 12:52:02 Magnet Current, Imag = 10.0106 A Vtrans = 0.100106 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 12:52:04 Magnet Current, Imag = 10.0108 A Vtrans = 0.100108 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 12:52:05 Magnet Current, Imag = 10.0110 A Vtrans = 0.100110 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 12:52:05 Magnet Current Measurement: Imag = 10.0110 +- 0.0003 A 12:52:05 Magnet current ramping to 15.0000 A... 12:52:15 Magnet Current, Imag = 14.9647 A Vtrans = 0.149647 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 12:52:17 Magnet Current, Imag = 14.9599 A Vtrans = 0.149599 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 12:52:19 Magnet Current, Imag = 14.9599 A Vtrans = 0.149599 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 12:52:20 Magnet Current, Imag = 14.9604 A Vtrans = 0.149604 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 12:52:20 Magnet Current Measurement: Imag = 14.9612 +- 0.0020 A 12:52:20 Magnetic Field Measurement Measuring the magnet strength... 12:52:20 Integrated Voltage Measurement... 12:52:43 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = 0.0100000 m VT = -0.003958802 Vs 12:52:58 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = -0.0100000 m VT = 0.004072511 Vs 12:53:13 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = 0.0100000 m VT = -0.003967948 Vs 12:53:28 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = -0.0100000 m VT = 0.004055660 Vs 12:53:43 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = 0.0100000 m VT = -0.003977011 Vs 12:53:58 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = -0.0100000 m VT = 0.004062672 Vs 12:54:13 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = 0.0100000 m VT = -0.003979640 Vs 12:54:28 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = -0.0100000 m VT = 0.004052012 Vs 12:54:28 Integrated Voltage Average and RMS of Measurements X0 = -0.0200000 m, Delta X = 0.0100000 m VT = -0.004015782 +- 0.000002843 Vs 12:54:28 Integrated Field Strength, Average and RMS of Measurements X0 = -0.0200000 m, dx = 0.0100000 m, Nturns = 10.0 VT = -0.004015782 +- 0.000002843 Vs BL = -0.04015782 +- 0.00002843 Tm 12:54:28 Integrated Voltage Measurement... 12:54:47 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = 0.0100000 m VT = -0.002040008 Vs 12:55:02 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = -0.0100000 m VT = 0.002072271 Vs 12:55:17 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = 0.0100000 m VT = -0.002059517 Vs 12:55:32 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = -0.0100000 m VT = 0.002066834 Vs 12:55:47 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = 0.0100000 m VT = -0.002073893 Vs 12:56:02 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = -0.0100000 m VT = 0.002038323 Vs 12:56:17 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = 0.0100000 m VT = -0.002094824 Vs 12:56:32 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = -0.0100000 m VT = 0.002024218 Vs 12:56:32 Integrated Voltage Average and RMS of Measurements X0 = -0.0100000 m, Delta X = 0.0100000 m VT = -0.002058736 +- 0.000002914 Vs 12:56:32 Integrated Field Strength, Average and RMS of Measurements X0 = -0.0100000 m, dx = 0.0100000 m, Nturns = 10.0 VT = -0.002058736 +- 0.000002914 Vs BL = -0.02058736 +- 0.00002914 Tm 12:56:32 Integrated Voltage Measurement... 12:56:51 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = -0.000166362 Vs 12:57:06 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = 0.000052580 Vs 12:57:21 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = -0.000184983 Vs 12:57:36 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = 0.000048425 Vs 12:57:51 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = -0.000180324 Vs 12:58:06 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = 0.000049102 Vs 12:58:21 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = -0.000182604 Vs 12:58:36 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = 0.000040686 Vs 12:58:36 Integrated Voltage Average and RMS of Measurements X0 = 0.0000000 m, Delta X = 0.0100000 m VT = -0.000113133 +- 0.000002778 Vs 12:58:36 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0000000 m, dx = 0.0100000 m, Nturns = 10.0 VT = -0.000113133 +- 0.000002778 Vs BL = -0.00113133 +- 0.00002778 Tm 12:58:36 Integrated Voltage Measurement... 12:58:55 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = 0.0100000 m VT = 0.001761258 Vs 12:59:10 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = -0.0100000 m VT = -0.001900945 Vs 12:59:25 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = 0.0100000 m VT = 0.001768374 Vs 12:59:40 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = -0.0100000 m VT = -0.001895061 Vs 12:59:55 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = 0.0100000 m VT = 0.001768519 Vs 13:00:10 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = -0.0100000 m VT = -0.001891830 Vs 13:00:25 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = 0.0100000 m VT = 0.001781141 Vs 13:00:40 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = -0.0100000 m VT = -0.001879702 Vs 13:00:40 Integrated Voltage Average and RMS of Measurements X0 = 0.0100000 m, Delta X = 0.0100000 m VT = 0.001830854 +- 0.000000603 Vs 13:00:40 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0100000 m, dx = 0.0100000 m, Nturns = 10.0 VT = 0.001830854 +- 0.000000603 Vs BL = 0.01830854 +- 0.00000603 Tm 13:00:40 Integrated Voltage Measurement... 13:00:59 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = 0.0100000 m VT = 0.003746757 Vs 13:01:14 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = -0.0100000 m VT = -0.003819513 Vs 13:01:29 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = 0.0100000 m VT = 0.003738675 Vs 13:01:44 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = -0.0100000 m VT = -0.003824364 Vs 13:01:59 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = 0.0100000 m VT = 0.003744126 Vs 13:02:14 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = -0.0100000 m VT = -0.003827498 Vs 13:02:30 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = 0.0100000 m VT = 0.003728612 Vs 13:02:45 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = -0.0100000 m VT = -0.003836540 Vs 13:02:45 Integrated Voltage Average and RMS of Measurements X0 = 0.0200000 m, Delta X = 0.0100000 m VT = 0.003783261 +- 0.000001583 Vs 13:02:45 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0200000 m, dx = 0.0100000 m, Nturns = 10.0 VT = 0.003783261 +- 0.000001583 Vs BL = 0.03783261 +- 0.00001583 Tm 13:02:50 BL vs X Fit Coefficients: Cn sigCn n (Tm/m^n) ----- --------+-------- 0 -0.00115 0.00001 1 1.94877 0.00072 Compute Fit Residuals: X0 BL sigBL BLfit BL-BLfit (m) (Tm) (Tm) (Tm) (Tm) ---------- --------+-------- -------- -------- -0.020000 -0.04016 0.00003 -0.04012 -0.00004 -0.010000 -0.02059 0.00003 -0.02063 0.00005 0.000000 -0.00113 0.00003 -0.00115 0.00002 0.010000 0.01831 0.00001 0.01834 -0.00003 0.020000 0.03783 0.00002 0.03783 0.00000 13:02:50 Magnet Strength Calculation Harmonic Number = 2 Number Of X0 Positions = 5 X0 BL sigBL (m) (Tm) (Tm) -------- --------+-------- -0.0200 -0.04016 0.00003 -0.0100 -0.02059 0.00003 0.0000 -0.00113 0.00003 0.0100 0.01831 0.00001 0.0200 0.03783 0.00002 SL = 1.9487675 +- 0.0007156 Tm^2/m^n 13:02:50 Quadrupole Strength Calculation Check Slope = 1.948768 Tm/m, Inter = -0.001147 Tm, Sig_slope = 0.001264 Tm/m, Est_sig = 0.000040 Tm GL = 1.948768 +- 0.001264 T 13:02:50 Magnet Strength Measurement Harmonic Number = 2 SL = 1.9487675 +- 0.0007156 Tm^2/m^n 13:02:52 Magnet Current, Imag = 14.9580 A Vtrans = 0.149580 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 13:02:54 Magnet Current, Imag = 14.9588 A Vtrans = 0.149588 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 13:02:55 Magnet Current, Imag = 14.9584 A Vtrans = 0.149584 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 13:02:57 Magnet Current, Imag = 14.9588 A Vtrans = 0.149588 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 13:02:57 Magnet Current Measurement: Imag = 14.9585 +- 0.0003 A 13:02:57 Magnet current ramping to 20.0000 A... 13:03:07 Magnet Current, Imag = 20.0594 A Vtrans = 0.200594 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 13:03:09 Magnet Current, Imag = 20.0534 A Vtrans = 0.200534 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 13:03:10 Magnet Current, Imag = 20.0537 A Vtrans = 0.200537 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 13:03:12 Magnet Current, Imag = 20.0538 A Vtrans = 0.200538 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 13:03:12 Magnet Current Measurement: Imag = 20.0551 +- 0.0025 A 13:03:12 Magnetic Field Measurement Measuring the magnet strength... 13:03:12 Integrated Voltage Measurement... 13:03:35 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = 0.0100000 m VT = -0.005423790 Vs 13:03:50 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = -0.0100000 m VT = 0.005333553 Vs 13:04:05 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = 0.0100000 m VT = -0.005414335 Vs 13:04:20 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = -0.0100000 m VT = 0.005339655 Vs 13:04:35 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = 0.0100000 m VT = -0.005425486 Vs 13:04:50 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = -0.0100000 m VT = 0.005317744 Vs 13:05:05 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = 0.0100000 m VT = -0.005457003 Vs 13:05:20 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = -0.0100000 m VT = 0.005301616 Vs 13:05:20 Integrated Voltage Average and RMS of Measurements X0 = -0.0200000 m, Delta X = 0.0100000 m VT = -0.005376648 +- 0.000003026 Vs 13:05:20 Integrated Field Strength, Average and RMS of Measurements X0 = -0.0200000 m, dx = 0.0100000 m, Nturns = 10.0 VT = -0.005376648 +- 0.000003026 Vs BL = -0.05376648 +- 0.00003026 Tm 13:05:20 Integrated Voltage Measurement... 13:05:39 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = 0.0100000 m VT = -0.002829964 Vs 13:05:54 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = -0.0100000 m VT = 0.002723009 Vs 13:06:09 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = 0.0100000 m VT = -0.002802303 Vs 13:06:24 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = -0.0100000 m VT = 0.002733560 Vs 13:06:39 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = 0.0100000 m VT = -0.002796888 Vs 13:06:54 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = -0.0100000 m VT = 0.002730561 Vs 13:07:09 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = 0.0100000 m VT = -0.002798067 Vs 13:07:24 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = -0.0100000 m VT = 0.002730685 Vs 13:07:25 Integrated Voltage Average and RMS of Measurements X0 = -0.0100000 m, Delta X = 0.0100000 m VT = -0.002768130 +- 0.000005084 Vs 13:07:25 Integrated Field Strength, Average and RMS of Measurements X0 = -0.0100000 m, dx = 0.0100000 m, Nturns = 10.0 VT = -0.002768130 +- 0.000005084 Vs BL = -0.02768130 +- 0.00005084 Tm 13:07:25 Integrated Voltage Measurement... 13:07:43 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = -0.000174239 Vs 13:07:58 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = 0.000127475 Vs 13:08:13 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = -0.000175690 Vs 13:08:28 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = 0.000132729 Vs 13:08:44 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = -0.000177743 Vs 13:08:59 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = 0.000115196 Vs 13:09:14 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = -0.000204137 Vs 13:09:29 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = 0.000108713 Vs 13:09:29 Integrated Voltage Average and RMS of Measurements X0 = 0.0000000 m, Delta X = 0.0100000 m VT = -0.000151990 +- 0.000003754 Vs 13:09:29 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0000000 m, dx = 0.0100000 m, Nturns = 10.0 VT = -0.000151990 +- 0.000003754 Vs BL = -0.00151990 +- 0.00003754 Tm 13:09:29 Integrated Voltage Measurement... 13:09:47 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = 0.0100000 m VT = 0.002407947 Vs 13:10:02 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = -0.0100000 m VT = -0.002517733 Vs 13:10:17 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = 0.0100000 m VT = 0.002402270 Vs 13:10:32 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = -0.0100000 m VT = -0.002501812 Vs 13:10:47 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = 0.0100000 m VT = 0.002423472 Vs 13:11:02 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = -0.0100000 m VT = -0.002489725 Vs 13:11:18 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = 0.0100000 m VT = 0.002425281 Vs 13:11:33 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = -0.0100000 m VT = -0.002478493 Vs 13:11:33 Integrated Voltage Average and RMS of Measurements X0 = 0.0100000 m, Delta X = 0.0100000 m VT = 0.002455842 +- 0.000004462 Vs 13:11:33 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0100000 m, dx = 0.0100000 m, Nturns = 10.0 VT = 0.002455842 +- 0.000004462 Vs BL = 0.02455842 +- 0.00004462 Tm 13:11:33 Integrated Voltage Measurement... 13:11:51 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = 0.0100000 m VT = 0.005056147 Vs 13:12:06 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = -0.0100000 m VT = -0.005080411 Vs 13:12:21 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = 0.0100000 m VT = 0.005043808 Vs 13:12:36 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = -0.0100000 m VT = -0.005094024 Vs 13:12:51 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = 0.0100000 m VT = 0.005052780 Vs 13:13:06 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = -0.0100000 m VT = -0.005073444 Vs 13:13:21 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = 0.0100000 m VT = 0.005070812 Vs 13:13:37 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = -0.0100000 m VT = -0.005052505 Vs 13:13:37 Integrated Voltage Average and RMS of Measurements X0 = 0.0200000 m, Delta X = 0.0100000 m VT = 0.005065491 +- 0.000003156 Vs 13:13:37 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0200000 m, dx = 0.0100000 m, Nturns = 10.0 VT = 0.005065491 +- 0.000003156 Vs BL = 0.05065491 +- 0.00003156 Tm 13:13:42 BL vs X Fit Coefficients: Cn sigCn n (Tm/m^n) ----- --------+-------- 0 -0.00155 0.00002 1 2.61082 0.00111 Compute Fit Residuals: X0 BL sigBL BLfit BL-BLfit (m) (Tm) (Tm) (Tm) (Tm) ---------- --------+-------- -------- -------- -0.020000 -0.05377 0.00003 -0.05377 0.00000 -0.010000 -0.02768 0.00005 -0.02766 -0.00002 0.000000 -0.00152 0.00004 -0.00155 0.00003 0.010000 0.02456 0.00004 0.02456 0.00000 0.020000 0.05065 0.00003 0.05067 -0.00001 13:13:42 Magnet Strength Calculation Harmonic Number = 2 Number Of X0 Positions = 5 X0 BL sigBL (m) (Tm) (Tm) -------- --------+-------- -0.0200 -0.05377 0.00003 -0.0100 -0.02768 0.00005 0.0000 -0.00152 0.00004 0.0100 0.02456 0.00004 0.0200 0.05065 0.00003 SL = 2.6108249 +- 0.0011056 Tm^2/m^n 13:13:42 Quadrupole Strength Calculation Check Slope = 2.610825 Tm/m, Inter = -0.001551 Tm, Sig_slope = 0.000723 Tm/m, Est_sig = 0.000023 Tm GL = 2.610825 +- 0.000723 T 13:13:42 Magnet Strength Measurement Harmonic Number = 2 SL = 2.6108249 +- 0.0011056 Tm^2/m^n 13:13:44 Magnet Current, Imag = 20.0542 A Vtrans = 0.200542 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 13:13:45 Magnet Current, Imag = 20.0545 A Vtrans = 0.200545 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 13:13:47 Magnet Current, Imag = 20.0543 A Vtrans = 0.200543 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 13:13:49 Magnet Current, Imag = 20.0548 A Vtrans = 0.200548 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 13:13:49 Magnet Current Measurement: Imag = 20.0545 +- 0.0002 A 13:13:49 Magnet current ramping to 25.0000 A... 13:13:59 Magnet Current, Imag = 25.0751 A Vtrans = 0.250751 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 13:14:00 Magnet Current, Imag = 25.0699 A Vtrans = 0.250699 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 13:14:02 Magnet Current, Imag = 25.0697 A Vtrans = 0.250697 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 13:14:04 Magnet Current, Imag = 25.0699 A Vtrans = 0.250699 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 13:14:04 Magnet Current Measurement: Imag = 25.0712 +- 0.0023 A 13:14:04 Magnetic Field Measurement Measuring the magnet strength... 13:14:04 Integrated Voltage Measurement... 13:14:26 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = 0.0100000 m VT = -0.006709737 Vs 13:14:41 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = -0.0100000 m VT = 0.006738449 Vs 13:14:56 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = 0.0100000 m VT = -0.006712358 Vs 13:15:12 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = -0.0100000 m VT = 0.006736313 Vs 13:15:27 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = 0.0100000 m VT = -0.006717717 Vs 13:15:42 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = -0.0100000 m VT = 0.006740424 Vs 13:15:57 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = 0.0100000 m VT = -0.006714281 Vs 13:16:12 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = -0.0100000 m VT = 0.006726759 Vs 13:16:12 Integrated Voltage Average and RMS of Measurements X0 = -0.0200000 m, Delta X = 0.0100000 m VT = -0.006724505 +- 0.000003038 Vs 13:16:12 Integrated Field Strength, Average and RMS of Measurements X0 = -0.0200000 m, dx = 0.0100000 m, Nturns = 10.0 VT = -0.006724505 +- 0.000003038 Vs BL = -0.06724505 +- 0.00003038 Tm 13:16:12 Integrated Voltage Measurement... 13:16:30 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = 0.0100000 m VT = -0.003464451 Vs 13:16:46 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = -0.0100000 m VT = 0.003447291 Vs 13:17:01 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = 0.0100000 m VT = -0.003465426 Vs 13:17:16 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = -0.0100000 m VT = 0.003443010 Vs 13:17:31 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = 0.0100000 m VT = -0.003461671 Vs 13:17:46 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = -0.0100000 m VT = 0.003451873 Vs 13:18:01 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = 0.0100000 m VT = -0.003461061 Vs 13:18:16 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = -0.0100000 m VT = 0.003460949 Vs 13:18:16 Integrated Voltage Average and RMS of Measurements X0 = -0.0100000 m, Delta X = 0.0100000 m VT = -0.003456967 +- 0.000002505 Vs 13:18:16 Integrated Field Strength, Average and RMS of Measurements X0 = -0.0100000 m, dx = 0.0100000 m, Nturns = 10.0 VT = -0.003456967 +- 0.000002505 Vs BL = -0.03456967 +- 0.00002505 Tm 13:18:16 Integrated Voltage Measurement... 13:18:35 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = -0.000199198 Vs 13:18:50 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = 0.000172707 Vs 13:19:05 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = -0.000197940 Vs 13:19:20 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = 0.000185465 Vs 13:19:35 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = -0.000198774 Vs 13:19:50 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = 0.000177504 Vs 13:20:05 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = -0.000198418 Vs 13:20:20 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = 0.000185442 Vs 13:20:20 Integrated Voltage Average and RMS of Measurements X0 = 0.0000000 m, Delta X = 0.0100000 m VT = -0.000189431 +- 0.000002509 Vs 13:20:20 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0000000 m, dx = 0.0100000 m, Nturns = 10.0 VT = -0.000189431 +- 0.000002509 Vs BL = -0.00189431 +- 0.00002509 Tm 13:20:20 Integrated Voltage Measurement... 13:20:39 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = 0.0100000 m VT = 0.003062414 Vs 13:20:54 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = -0.0100000 m VT = -0.003091657 Vs 13:21:09 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = 0.0100000 m VT = 0.003062182 Vs 13:21:24 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = -0.0100000 m VT = -0.003087493 Vs 13:21:39 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = 0.0100000 m VT = 0.003069069 Vs 13:21:54 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = -0.0100000 m VT = -0.003077270 Vs 13:22:09 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = 0.0100000 m VT = 0.003071461 Vs 13:22:25 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = -0.0100000 m VT = -0.003080532 Vs 13:22:25 Integrated Voltage Average and RMS of Measurements X0 = 0.0100000 m, Delta X = 0.0100000 m VT = 0.003075260 +- 0.000001436 Vs 13:22:25 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0100000 m, dx = 0.0100000 m, Nturns = 10.0 VT = 0.003075260 +- 0.000001436 Vs BL = 0.03075260 +- 0.00001436 Tm 13:22:25 Integrated Voltage Measurement... 13:22:43 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = 0.0100000 m VT = 0.006348113 Vs 13:22:58 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = -0.0100000 m VT = -0.006324938 Vs 13:23:13 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = 0.0100000 m VT = 0.006355916 Vs 13:23:28 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = -0.0100000 m VT = -0.006328772 Vs 13:23:43 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = 0.0100000 m VT = 0.006351227 Vs 13:23:58 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = -0.0100000 m VT = -0.006326228 Vs 13:24:13 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = 0.0100000 m VT = 0.006353810 Vs 13:24:29 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = -0.0100000 m VT = -0.006326208 Vs 13:24:29 Integrated Voltage Average and RMS of Measurements X0 = 0.0200000 m, Delta X = 0.0100000 m VT = 0.006339402 +- 0.000002107 Vs 13:24:29 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0200000 m, dx = 0.0100000 m, Nturns = 10.0 VT = 0.006339402 +- 0.000002107 Vs BL = 0.06339402 +- 0.00002107 Tm 13:24:34 BL vs X Fit Coefficients: Cn sigCn n (Tm/m^n) ----- --------+-------- 0 -0.00191 0.00001 1 3.26600 0.00079 Compute Fit Residuals: X0 BL sigBL BLfit BL-BLfit (m) (Tm) (Tm) (Tm) (Tm) ---------- --------+-------- -------- -------- -0.020000 -0.06725 0.00003 -0.06723 -0.00001 -0.010000 -0.03457 0.00003 -0.03457 0.00000 0.000000 -0.00189 0.00003 -0.00191 0.00002 0.010000 0.03075 0.00001 0.03075 0.00001 0.020000 0.06339 0.00002 0.06341 -0.00001 13:24:34 Magnet Strength Calculation Harmonic Number = 2 Number Of X0 Positions = 5 X0 BL sigBL (m) (Tm) (Tm) -------- --------+-------- -0.0200 -0.06725 0.00003 -0.0100 -0.03457 0.00003 0.0000 -0.00189 0.00003 0.0100 0.03075 0.00001 0.0200 0.06339 0.00002 SL = 3.2660040 +- 0.0007938 Tm^2/m^n 13:24:34 Quadrupole Strength Calculation Check Slope = 3.266004 Tm/m, Inter = -0.001912 Tm, Sig_slope = 0.000484 Tm/m, Est_sig = 0.000015 Tm GL = 3.266004 +- 0.000484 T 13:24:34 Magnet Strength Measurement Harmonic Number = 2 SL = 3.2660040 +- 0.0007938 Tm^2/m^n 13:24:36 Magnet Current, Imag = 25.0707 A Vtrans = 0.250707 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 13:24:37 Magnet Current, Imag = 25.0702 A Vtrans = 0.250702 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 13:24:39 Magnet Current, Imag = 25.0699 A Vtrans = 0.250699 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 13:24:41 Magnet Current, Imag = 25.0701 A Vtrans = 0.250701 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 13:24:41 Magnet Current Measurement: Imag = 25.0702 +- 0.0003 A 13:24:41 Magnet current ramping to 30.0000 A... 13:24:50 Magnet Current, Imag = 30.1051 A Vtrans = 0.301051 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 13:24:52 Magnet Current, Imag = 30.1034 A Vtrans = 0.301034 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 13:24:53 Magnet Current, Imag = 30.1009 A Vtrans = 0.301009 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 13:24:54 Magnet Current, Imag = 30.1017 A Vtrans = 0.301017 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 13:24:54 Magnet Current Measurement: Imag = 30.1028 +- 0.0016 A 13:24:54 Magnetic Field Measurement Measuring the magnet strength... 13:24:54 Integrated Voltage Measurement... 13:25:16 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = 0.0100000 m VT = -0.008075048 Vs 13:25:32 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = -0.0100000 m VT = 0.008082187 Vs 13:25:47 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = 0.0100000 m VT = -0.008065184 Vs 13:26:02 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = -0.0100000 m VT = 0.008080582 Vs 13:26:17 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = 0.0100000 m VT = -0.008058672 Vs 13:26:32 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = -0.0100000 m VT = 0.008089895 Vs 13:26:47 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = 0.0100000 m VT = -0.008063628 Vs 13:27:02 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = -0.0100000 m VT = 0.008081497 Vs 13:27:02 Integrated Voltage Average and RMS of Measurements X0 = -0.0200000 m, Delta X = 0.0100000 m VT = -0.008074587 +- 0.000002416 Vs 13:27:02 Integrated Field Strength, Average and RMS of Measurements X0 = -0.0200000 m, dx = 0.0100000 m, Nturns = 10.0 VT = -0.008074587 +- 0.000002416 Vs BL = -0.08074587 +- 0.00002416 Tm 13:27:02 Integrated Voltage Measurement... 13:27:20 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = 0.0100000 m VT = -0.004152725 Vs 13:27:35 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = -0.0100000 m VT = 0.004139190 Vs 13:27:51 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = 0.0100000 m VT = -0.004146762 Vs 13:28:06 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = -0.0100000 m VT = 0.004151930 Vs 13:28:21 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = 0.0100000 m VT = -0.004152622 Vs 13:28:36 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = -0.0100000 m VT = 0.004150819 Vs 13:28:51 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = 0.0100000 m VT = -0.004143735 Vs 13:29:06 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = -0.0100000 m VT = 0.004151693 Vs 13:29:06 Integrated Voltage Average and RMS of Measurements X0 = -0.0100000 m, Delta X = 0.0100000 m VT = -0.004148684 +- 0.000002123 Vs 13:29:06 Integrated Field Strength, Average and RMS of Measurements X0 = -0.0100000 m, dx = 0.0100000 m, Nturns = 10.0 VT = -0.004148684 +- 0.000002123 Vs BL = -0.04148684 +- 0.00002123 Tm 13:29:06 Integrated Voltage Measurement... 13:29:24 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = -0.000223111 Vs 13:29:39 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = 0.000223456 Vs 13:29:54 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = -0.000231595 Vs 13:30:10 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = 0.000211975 Vs 13:30:25 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = -0.000237406 Vs 13:30:40 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = 0.000215321 Vs 13:30:55 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = -0.000229015 Vs 13:31:10 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = 0.000220063 Vs 13:31:10 Integrated Voltage Average and RMS of Measurements X0 = 0.0000000 m, Delta X = 0.0100000 m VT = -0.000223993 +- 0.000001681 Vs 13:31:10 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0000000 m, dx = 0.0100000 m, Nturns = 10.0 VT = -0.000223993 +- 0.000001681 Vs BL = -0.00223993 +- 0.00001681 Tm 13:31:10 Integrated Voltage Measurement... 13:31:29 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = 0.0100000 m VT = 0.003710165 Vs 13:31:44 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = -0.0100000 m VT = -0.003685097 Vs 13:31:59 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = 0.0100000 m VT = 0.003698876 Vs 13:32:14 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = -0.0100000 m VT = -0.003700504 Vs 13:32:29 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = 0.0100000 m VT = 0.003687128 Vs 13:32:44 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = -0.0100000 m VT = -0.003701676 Vs 13:33:00 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = 0.0100000 m VT = 0.003698995 Vs 13:33:15 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = -0.0100000 m VT = -0.003694388 Vs 13:33:15 Integrated Voltage Average and RMS of Measurements X0 = 0.0100000 m, Delta X = 0.0100000 m VT = 0.003697104 +- 0.000001900 Vs 13:33:15 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0100000 m, dx = 0.0100000 m, Nturns = 10.0 VT = 0.003697104 +- 0.000001900 Vs BL = 0.03697104 +- 0.00001900 Tm 13:33:15 Integrated Voltage Measurement... 13:33:33 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = 0.0100000 m VT = 0.007602648 Vs 13:33:48 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = -0.0100000 m VT = -0.007628802 Vs 13:34:03 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = 0.0100000 m VT = 0.007593997 Vs 13:34:19 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = -0.0100000 m VT = -0.007631875 Vs 13:34:34 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = 0.0100000 m VT = 0.007595705 Vs 13:34:49 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = -0.0100000 m VT = -0.007623469 Vs 13:35:04 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = 0.0100000 m VT = 0.007617268 Vs 13:35:19 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = -0.0100000 m VT = -0.007619108 Vs 13:35:19 Integrated Voltage Average and RMS of Measurements X0 = 0.0200000 m, Delta X = 0.0100000 m VT = 0.007614109 +- 0.000003204 Vs 13:35:19 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0200000 m, dx = 0.0100000 m, Nturns = 10.0 VT = 0.007614109 +- 0.000003204 Vs BL = 0.07614109 +- 0.00003204 Tm 13:35:24 BL vs X Fit Coefficients: Cn sigCn n (Tm/m^n) ----- --------+-------- 0 -0.00227 0.00001 1 3.92232 0.00085 Compute Fit Residuals: X0 BL sigBL BLfit BL-BLfit (m) (Tm) (Tm) (Tm) (Tm) ---------- --------+-------- -------- -------- -0.020000 -0.08075 0.00002 -0.08072 -0.00003 -0.010000 -0.04149 0.00002 -0.04150 0.00001 0.000000 -0.00224 0.00002 -0.00227 0.00003 0.010000 0.03697 0.00002 0.03695 0.00002 0.020000 0.07614 0.00003 0.07617 -0.00003 13:35:24 Magnet Strength Calculation Harmonic Number = 2 Number Of X0 Positions = 5 X0 BL sigBL (m) (Tm) (Tm) -------- --------+-------- -0.0200 -0.08075 0.00002 -0.0100 -0.04149 0.00002 0.0000 -0.00224 0.00002 0.0100 0.03697 0.00002 0.0200 0.07614 0.00003 SL = 3.9223180 +- 0.0008516 Tm^2/m^n 13:35:24 Quadrupole Strength Calculation Check Slope = 3.922318 Tm/m, Inter = -0.002272 Tm, Sig_slope = 0.001058 Tm/m, Est_sig = 0.000033 Tm GL = 3.922318 +- 0.001058 T 13:35:24 Magnet Strength Measurement Harmonic Number = 2 SL = 3.9223180 +- 0.0008516 Tm^2/m^n 13:35:25 Magnet Current, Imag = 30.1023 A Vtrans = 0.301023 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 13:35:27 Magnet Current, Imag = 30.1020 A Vtrans = 0.301020 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 13:35:28 Magnet Current, Imag = 30.1029 A Vtrans = 0.301029 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 13:35:29 Magnet Current, Imag = 30.1023 A Vtrans = 0.301023 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 13:35:29 Magnet Current Measurement: Imag = 30.1024 +- 0.0003 A 13:35:30 Magnet current ramping to 35.0000 A... 13:35:38 Magnet Current, Imag = 35.0852 A Vtrans = 0.350852 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 13:35:39 Magnet Current, Imag = 35.0840 A Vtrans = 0.350840 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 13:35:41 Magnet Current, Imag = 35.0815 A Vtrans = 0.350815 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 13:35:42 Magnet Current, Imag = 35.0817 A Vtrans = 0.350817 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 13:35:42 Magnet Current Measurement: Imag = 35.0831 +- 0.0016 A 13:35:42 Magnetic Field Measurement Measuring the magnet strength... 13:35:42 Integrated Voltage Measurement... 13:36:04 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = 0.0100000 m VT = -0.009410846 Vs 13:36:19 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = -0.0100000 m VT = 0.009411555 Vs 13:36:34 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = 0.0100000 m VT = -0.009415241 Vs 13:36:50 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = -0.0100000 m VT = 0.009408601 Vs 13:37:05 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = 0.0100000 m VT = -0.009410559 Vs 13:37:20 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = -0.0100000 m VT = 0.009403359 Vs 13:37:35 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = 0.0100000 m VT = -0.009424417 Vs 13:37:50 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = -0.0100000 m VT = 0.009389915 Vs 13:37:50 Integrated Voltage Average and RMS of Measurements X0 = -0.0200000 m, Delta X = 0.0100000 m VT = -0.009409312 +- 0.000002265 Vs 13:37:50 Integrated Field Strength, Average and RMS of Measurements X0 = -0.0200000 m, dx = 0.0100000 m, Nturns = 10.0 VT = -0.009409312 +- 0.000002265 Vs BL = -0.09409312 +- 0.00002265 Tm 13:37:50 Integrated Voltage Measurement... 13:38:09 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = 0.0100000 m VT = -0.004848231 Vs 13:38:24 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = -0.0100000 m VT = 0.004829771 Vs 13:38:39 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = 0.0100000 m VT = -0.004824101 Vs 13:38:54 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = -0.0100000 m VT = 0.004852389 Vs 13:39:09 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = 0.0100000 m VT = -0.004812359 Vs 13:39:24 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = -0.0100000 m VT = 0.004854316 Vs 13:39:39 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = 0.0100000 m VT = -0.004814981 Vs 13:39:54 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = -0.0100000 m VT = 0.004848341 Vs 13:39:55 Integrated Voltage Average and RMS of Measurements X0 = -0.0100000 m, Delta X = 0.0100000 m VT = -0.004835561 +- 0.000003130 Vs 13:39:55 Integrated Field Strength, Average and RMS of Measurements X0 = -0.0100000 m, dx = 0.0100000 m, Nturns = 10.0 VT = -0.004835561 +- 0.000003130 Vs BL = -0.04835561 +- 0.00003130 Tm 13:39:55 Integrated Voltage Measurement... 13:40:13 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = -0.000242551 Vs 13:40:28 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = 0.000276195 Vs 13:40:43 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = -0.000254931 Vs 13:40:58 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = 0.000259738 Vs 13:41:13 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = -0.000266075 Vs 13:41:29 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = 0.000254494 Vs 13:41:44 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = -0.000275066 Vs 13:41:59 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = 0.000250257 Vs 13:41:59 Integrated Voltage Average and RMS of Measurements X0 = 0.0000000 m, Delta X = 0.0100000 m VT = -0.000259913 +- 0.000001913 Vs 13:41:59 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0000000 m, dx = 0.0100000 m, Nturns = 10.0 VT = -0.000259913 +- 0.000001913 Vs BL = -0.00259913 +- 0.00001913 Tm 13:41:59 Integrated Voltage Measurement... 13:42:17 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = 0.0100000 m VT = 0.004295605 Vs 13:42:32 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = -0.0100000 m VT = -0.004320683 Vs 13:42:47 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = 0.0100000 m VT = 0.004291099 Vs 13:43:02 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = -0.0100000 m VT = -0.004313981 Vs 13:43:17 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = 0.0100000 m VT = 0.004304760 Vs 13:43:32 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = -0.0100000 m VT = -0.004315214 Vs 13:43:48 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = 0.0100000 m VT = 0.004301338 Vs 13:44:03 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = -0.0100000 m VT = -0.004311223 Vs 13:44:03 Integrated Voltage Average and RMS of Measurements X0 = 0.0100000 m, Delta X = 0.0100000 m VT = 0.004306738 +- 0.000002755 Vs 13:44:03 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0100000 m, dx = 0.0100000 m, Nturns = 10.0 VT = 0.004306738 +- 0.000002755 Vs BL = 0.04306738 +- 0.00002755 Tm 13:44:03 Integrated Voltage Measurement... 13:44:21 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = 0.0100000 m VT = 0.008870511 Vs 13:44:36 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = -0.0100000 m VT = -0.008887030 Vs 13:44:51 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = 0.0100000 m VT = 0.008862861 Vs 13:45:06 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = -0.0100000 m VT = -0.008879782 Vs 13:45:22 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = 0.0100000 m VT = 0.008866952 Vs 13:45:37 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = -0.0100000 m VT = -0.008888517 Vs 13:45:52 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = 0.0100000 m VT = 0.008863113 Vs 13:46:07 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = -0.0100000 m VT = -0.008886813 Vs 13:46:07 Integrated Voltage Average and RMS of Measurements X0 = 0.0200000 m, Delta X = 0.0100000 m VT = 0.008875697 +- 0.000002885 Vs 13:46:07 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0200000 m, dx = 0.0100000 m, Nturns = 10.0 VT = 0.008875697 +- 0.000002885 Vs BL = 0.08875697 +- 0.00002885 Tm 13:46:12 BL vs X Fit Coefficients: Cn sigCn n (Tm/m^n) ----- --------+-------- 0 -0.00264 0.00001 1 4.57123 0.00084 Compute Fit Residuals: X0 BL sigBL BLfit BL-BLfit (m) (Tm) (Tm) (Tm) (Tm) ---------- --------+-------- -------- -------- -0.020000 -0.09409 0.00002 -0.09407 -0.00002 -0.010000 -0.04836 0.00003 -0.04836 0.00000 0.000000 -0.00260 0.00002 -0.00264 0.00005 0.010000 0.04307 0.00003 0.04307 -0.00000 0.020000 0.08876 0.00003 0.08878 -0.00002 13:46:12 Magnet Strength Calculation Harmonic Number = 2 Number Of X0 Positions = 5 X0 BL sigBL (m) (Tm) (Tm) -------- --------+-------- -0.0200 -0.09409 0.00002 -0.0100 -0.04836 0.00003 0.0000 -0.00260 0.00002 0.0100 0.04307 0.00003 0.0200 0.08876 0.00003 SL = 4.5712317 +- 0.0008438 Tm^2/m^n 13:46:12 Quadrupole Strength Calculation Check Slope = 4.571232 Tm/m, Inter = -0.002645 Tm, Sig_slope = 0.001028 Tm/m, Est_sig = 0.000033 Tm GL = 4.571232 +- 0.001028 T 13:46:12 Magnet Strength Measurement Harmonic Number = 2 SL = 4.5712317 +- 0.0008438 Tm^2/m^n 13:46:13 Magnet Current, Imag = 35.0830 A Vtrans = 0.350830 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 13:46:15 Magnet Current, Imag = 35.0831 A Vtrans = 0.350831 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 13:46:16 Magnet Current, Imag = 35.0829 A Vtrans = 0.350829 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 13:46:17 Magnet Current, Imag = 35.0824 A Vtrans = 0.350824 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 13:46:17 Magnet Current Measurement: Imag = 35.0829 +- 0.0003 A 13:46:17 Magnet current ramping to 40.0000 A... 13:46:26 Magnet Current, Imag = 40.1692 A Vtrans = 0.401692 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 13:46:27 Magnet Current, Imag = 40.1687 A Vtrans = 0.401687 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 13:46:29 Magnet Current, Imag = 40.1662 A Vtrans = 0.401662 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 13:46:30 Magnet Current, Imag = 40.1663 A Vtrans = 0.401663 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 13:46:30 Magnet Current Measurement: Imag = 40.1676 +- 0.0014 A 13:46:30 Magnetic Field Measurement Measuring the magnet strength... 13:46:30 Integrated Voltage Measurement... 13:46:52 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = 0.0100000 m VT = -0.010778989 Vs 13:47:07 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = -0.0100000 m VT = 0.010767515 Vs 13:47:22 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = 0.0100000 m VT = -0.010782911 Vs 13:47:37 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = -0.0100000 m VT = 0.010758886 Vs 13:47:52 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = 0.0100000 m VT = -0.010796382 Vs 13:48:08 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = -0.0100000 m VT = 0.010752171 Vs 13:48:23 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = 0.0100000 m VT = -0.010796966 Vs 13:48:38 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = -0.0100000 m VT = 0.010750451 Vs 13:48:38 Integrated Voltage Average and RMS of Measurements X0 = -0.0200000 m, Delta X = 0.0100000 m VT = -0.010773034 +- 0.000001285 Vs 13:48:38 Integrated Field Strength, Average and RMS of Measurements X0 = -0.0200000 m, dx = 0.0100000 m, Nturns = 10.0 VT = -0.010773034 +- 0.000001285 Vs BL = -0.10773034 +- 0.00001285 Tm 13:48:38 Integrated Voltage Measurement... 13:48:57 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = 0.0100000 m VT = -0.005549282 Vs 13:49:12 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = -0.0100000 m VT = 0.005519128 Vs 13:49:27 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = 0.0100000 m VT = -0.005551996 Vs 13:49:42 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = -0.0100000 m VT = 0.005520730 Vs 13:49:57 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = 0.0100000 m VT = -0.005547160 Vs 13:50:12 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = -0.0100000 m VT = 0.005515956 Vs 13:50:27 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = 0.0100000 m VT = -0.005564092 Vs 13:50:43 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = -0.0100000 m VT = 0.005515650 Vs 13:50:43 Integrated Voltage Average and RMS of Measurements X0 = -0.0100000 m, Delta X = 0.0100000 m VT = -0.005535499 +- 0.000003044 Vs 13:50:43 Integrated Field Strength, Average and RMS of Measurements X0 = -0.0100000 m, dx = 0.0100000 m, Nturns = 10.0 VT = -0.005535499 +- 0.000003044 Vs BL = -0.05535499 +- 0.00003044 Tm 13:50:43 Integrated Voltage Measurement... 13:51:01 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = -0.000315964 Vs 13:51:16 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = 0.000287463 Vs 13:51:32 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = -0.000314095 Vs 13:51:47 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = 0.000284528 Vs 13:52:02 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = -0.000312317 Vs 13:52:17 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = 0.000282853 Vs 13:52:32 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = -0.000314525 Vs 13:52:47 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = 0.000273824 Vs 13:52:47 Integrated Voltage Average and RMS of Measurements X0 = 0.0000000 m, Delta X = 0.0100000 m VT = -0.000298196 +- 0.000002746 Vs 13:52:47 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0000000 m, dx = 0.0100000 m, Nturns = 10.0 VT = -0.000298196 +- 0.000002746 Vs BL = -0.00298196 +- 0.00002746 Tm 13:52:47 Integrated Voltage Measurement... 13:53:06 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = 0.0100000 m VT = 0.004908782 Vs 13:53:21 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = -0.0100000 m VT = -0.004958283 Vs 13:53:36 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = 0.0100000 m VT = 0.004914210 Vs 13:53:51 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = -0.0100000 m VT = -0.004947762 Vs 13:54:06 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = 0.0100000 m VT = 0.004933319 Vs 13:54:21 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = -0.0100000 m VT = -0.004935405 Vs 13:54:36 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = 0.0100000 m VT = 0.004925480 Vs 13:54:51 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = -0.0100000 m VT = -0.004936477 Vs 13:54:51 Integrated Voltage Average and RMS of Measurements X0 = 0.0100000 m, Delta X = 0.0100000 m VT = 0.004932465 +- 0.000001511 Vs 13:54:51 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0100000 m, dx = 0.0100000 m, Nturns = 10.0 VT = 0.004932465 +- 0.000001511 Vs BL = 0.04932465 +- 0.00001511 Tm 13:54:51 Integrated Voltage Measurement... 13:55:10 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = 0.0100000 m VT = 0.010150460 Vs 13:55:25 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = -0.0100000 m VT = -0.010181521 Vs 13:55:40 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = 0.0100000 m VT = 0.010162495 Vs 13:55:55 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = -0.0100000 m VT = -0.010167211 Vs 13:56:10 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = 0.0100000 m VT = 0.010146130 Vs 13:56:25 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = -0.0100000 m VT = -0.010198113 Vs 13:56:40 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = 0.0100000 m VT = 0.010125432 Vs 13:56:55 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = -0.0100000 m VT = -0.010193483 Vs 13:56:55 Integrated Voltage Average and RMS of Measurements X0 = 0.0200000 m, Delta X = 0.0100000 m VT = 0.010165606 +- 0.000004499 Vs 13:56:55 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0200000 m, dx = 0.0100000 m, Nturns = 10.0 VT = 0.010165606 +- 0.000004499 Vs BL = 0.10165606 +- 0.00004499 Tm 13:57:01 BL vs X Fit Coefficients: Cn sigCn n (Tm/m^n) ----- --------+-------- 0 -0.00302 0.00001 1 5.23452 0.00100 Compute Fit Residuals: X0 BL sigBL BLfit BL-BLfit (m) (Tm) (Tm) (Tm) (Tm) ---------- --------+-------- -------- -------- -0.020000 -0.10773 0.00001 -0.10771 -0.00002 -0.010000 -0.05535 0.00003 -0.05536 0.00001 0.000000 -0.00298 0.00003 -0.00302 0.00004 0.010000 0.04932 0.00002 0.04933 -0.00000 0.020000 0.10166 0.00004 0.10167 -0.00002 13:57:01 Magnet Strength Calculation Harmonic Number = 2 Number Of X0 Positions = 5 X0 BL sigBL (m) (Tm) (Tm) -------- --------+-------- -0.0200 -0.10773 0.00001 -0.0100 -0.05535 0.00003 0.0000 -0.00298 0.00003 0.0100 0.04932 0.00002 0.0200 0.10166 0.00004 SL = 5.2345243 +- 0.0009956 Tm^2/m^n 13:57:01 Quadrupole Strength Calculation Check Slope = 5.234524 Tm/m, Inter = -0.003017 Tm, Sig_slope = 0.000840 Tm/m, Est_sig = 0.000027 Tm GL = 5.234524 +- 0.000840 T 13:57:01 Magnet Strength Measurement Harmonic Number = 2 SL = 5.2345243 +- 0.0009956 Tm^2/m^n 13:57:02 Magnet Current, Imag = 40.1676 A Vtrans = 0.401676 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 13:57:03 Magnet Current, Imag = 40.1679 A Vtrans = 0.401679 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 13:57:05 Magnet Current, Imag = 40.1678 A Vtrans = 0.401678 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 13:57:06 Magnet Current, Imag = 40.1674 A Vtrans = 0.401674 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 13:57:06 Magnet Current Measurement: Imag = 40.1677 +- 0.0002 A 13:57:06 Magnet current ramping to 45.0000 A... 13:57:15 Magnet Current, Imag = 45.2553 A Vtrans = 0.452553 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 13:57:16 Magnet Current, Imag = 45.2555 A Vtrans = 0.452555 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 13:57:17 Magnet Current, Imag = 45.2532 A Vtrans = 0.452532 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 13:57:19 Magnet Current, Imag = 45.2533 A Vtrans = 0.452533 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 13:57:19 Magnet Current Measurement: Imag = 45.2543 +- 0.0011 A 13:57:19 Magnetic Field Measurement Measuring the magnet strength... 13:57:19 Integrated Voltage Measurement... 13:57:41 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = 0.0100000 m VT = -0.012157643 Vs 13:57:56 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = -0.0100000 m VT = 0.012114391 Vs 13:58:12 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = 0.0100000 m VT = -0.012169581 Vs 13:58:27 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = -0.0100000 m VT = 0.012097150 Vs 13:58:42 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = 0.0100000 m VT = -0.012163751 Vs 13:58:57 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = -0.0100000 m VT = 0.012107611 Vs 13:59:12 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = 0.0100000 m VT = -0.012152765 Vs 13:59:27 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = -0.0100000 m VT = 0.012140239 Vs 13:59:27 Integrated Voltage Average and RMS of Measurements X0 = -0.0200000 m, Delta X = 0.0100000 m VT = -0.012137891 +- 0.000005075 Vs 13:59:27 Integrated Field Strength, Average and RMS of Measurements X0 = -0.0200000 m, dx = 0.0100000 m, Nturns = 10.0 VT = -0.012137891 +- 0.000005075 Vs BL = -0.12137891 +- 0.00005075 Tm 13:59:27 Integrated Voltage Measurement... 13:59:45 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = 0.0100000 m VT = -0.006238028 Vs 14:00:00 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = -0.0100000 m VT = 0.006230282 Vs 14:00:15 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = 0.0100000 m VT = -0.006234747 Vs 14:00:30 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = -0.0100000 m VT = 0.006231429 Vs 14:00:46 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = 0.0100000 m VT = -0.006233944 Vs 14:01:01 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = -0.0100000 m VT = 0.006247248 Vs 14:01:16 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = 0.0100000 m VT = -0.006214722 Vs 14:01:31 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = -0.0100000 m VT = 0.006244787 Vs 14:01:31 Integrated Voltage Average and RMS of Measurements X0 = -0.0100000 m, Delta X = 0.0100000 m VT = -0.006234398 +- 0.000003929 Vs 14:01:31 Integrated Field Strength, Average and RMS of Measurements X0 = -0.0100000 m, dx = 0.0100000 m, Nturns = 10.0 VT = -0.006234398 +- 0.000003929 Vs BL = -0.06234398 +- 0.00003929 Tm 14:01:31 Integrated Voltage Measurement... 14:01:49 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = -0.000327093 Vs 14:02:04 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = 0.000341632 Vs 14:02:19 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = -0.000335311 Vs 14:02:34 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = 0.000330645 Vs 14:02:49 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = -0.000334061 Vs 14:03:04 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = 0.000334821 Vs 14:03:19 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = -0.000334605 Vs 14:03:35 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = 0.000338077 Vs 14:03:35 Integrated Voltage Average and RMS of Measurements X0 = 0.0000000 m, Delta X = 0.0100000 m VT = -0.000334531 +- 0.000001196 Vs 14:03:35 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0000000 m, dx = 0.0100000 m, Nturns = 10.0 VT = -0.000334531 +- 0.000001196 Vs BL = -0.00334531 +- 0.00001196 Tm 14:03:35 Integrated Voltage Measurement... 14:03:53 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = 0.0100000 m VT = 0.005562299 Vs 14:04:08 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = -0.0100000 m VT = -0.005547214 Vs 14:04:23 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = 0.0100000 m VT = 0.005563883 Vs 14:04:38 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = -0.0100000 m VT = -0.005550765 Vs 14:04:53 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = 0.0100000 m VT = 0.005560211 Vs 14:05:08 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = -0.0100000 m VT = -0.005553458 Vs 14:05:23 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = 0.0100000 m VT = 0.005570138 Vs 14:05:39 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = -0.0100000 m VT = -0.005547607 Vs 14:05:39 Integrated Voltage Average and RMS of Measurements X0 = 0.0100000 m, Delta X = 0.0100000 m VT = 0.005556947 +- 0.000001471 Vs 14:05:39 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0100000 m, dx = 0.0100000 m, Nturns = 10.0 VT = 0.005556947 +- 0.000001471 Vs BL = 0.05556947 +- 0.00001471 Tm 14:05:39 Integrated Voltage Measurement... 14:05:58 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = 0.0100000 m VT = 0.011455759 Vs 14:06:13 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = -0.0100000 m VT = -0.011455468 Vs 14:06:28 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = 0.0100000 m VT = 0.011430133 Vs 14:06:43 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = -0.0100000 m VT = -0.011471825 Vs 14:06:58 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = 0.0100000 m VT = 0.011444978 Vs 14:07:13 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = -0.0100000 m VT = -0.011452024 Vs 14:07:28 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = 0.0100000 m VT = 0.011448059 Vs 14:07:43 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = -0.0100000 m VT = -0.011450781 Vs 14:07:43 Integrated Voltage Average and RMS of Measurements X0 = 0.0200000 m, Delta X = 0.0100000 m VT = 0.011451128 +- 0.000002737 Vs 14:07:43 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0200000 m, dx = 0.0100000 m, Nturns = 10.0 VT = 0.011451128 +- 0.000002737 Vs BL = 0.11451128 +- 0.00002737 Tm 14:07:48 BL vs X Fit Coefficients: Cn sigCn n (Tm/m^n) ----- --------+-------- 0 -0.00340 0.00001 1 5.89694 0.00123 Compute Fit Residuals: X0 BL sigBL BLfit BL-BLfit (m) (Tm) (Tm) (Tm) (Tm) ---------- --------+-------- -------- -------- -0.020000 -0.12138 0.00005 -0.12134 -0.00004 -0.010000 -0.06234 0.00004 -0.06237 0.00002 0.000000 -0.00335 0.00001 -0.00340 0.00005 0.010000 0.05557 0.00001 0.05557 -0.00000 0.020000 0.11451 0.00003 0.11454 -0.00003 14:07:48 Magnet Strength Calculation Harmonic Number = 2 Number Of X0 Positions = 5 X0 BL sigBL (m) (Tm) (Tm) -------- --------+-------- -0.0200 -0.12138 0.00005 -0.0100 -0.06234 0.00004 0.0000 -0.00335 0.00001 0.0100 0.05557 0.00001 0.0200 0.11451 0.00003 SL = 5.8969385 +- 0.0012271 Tm^2/m^n 14:07:48 Quadrupole Strength Calculation Check Slope = 5.896938 Tm/m, Inter = -0.003397 Tm, Sig_slope = 0.001411 Tm/m, Est_sig = 0.000045 Tm GL = 5.896938 +- 0.001411 T 14:07:48 Magnet Strength Measurement Harmonic Number = 2 SL = 5.8969385 +- 0.0012271 Tm^2/m^n 14:07:50 Magnet Current, Imag = 45.2525 A Vtrans = 0.452525 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 14:07:51 Magnet Current, Imag = 45.2522 A Vtrans = 0.452522 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 14:07:52 Magnet Current, Imag = 45.2510 A Vtrans = 0.452510 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 14:07:53 Magnet Current, Imag = 45.2517 A Vtrans = 0.452517 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 14:07:53 Magnet Current Measurement: Imag = 45.2518 +- 0.0006 A 14:07:54 Magnet current ramping to 50.0000 A... 14:08:02 Magnet Current, Imag = 50.2692 A Vtrans = 0.502692 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 14:08:03 Magnet Current, Imag = 50.2694 A Vtrans = 0.502694 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 14:08:05 Magnet Current, Imag = 50.2673 A Vtrans = 0.502673 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 14:08:06 Magnet Current, Imag = 50.2679 A Vtrans = 0.502679 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 14:08:06 Magnet Current Measurement: Imag = 50.2685 +- 0.0009 A 14:08:06 Magnetic Field Measurement Measuring the magnet strength... 14:08:06 Integrated Voltage Measurement... 14:08:28 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = 0.0100000 m VT = -0.013462606 Vs 14:08:43 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = -0.0100000 m VT = 0.013485301 Vs 14:08:58 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = 0.0100000 m VT = -0.013469568 Vs 14:09:13 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = -0.0100000 m VT = 0.013487663 Vs 14:09:29 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = 0.0100000 m VT = -0.013457670 Vs 14:09:44 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = -0.0100000 m VT = 0.013495374 Vs 14:09:59 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = 0.0100000 m VT = -0.013471096 Vs 14:10:14 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = -0.0100000 m VT = 0.013486821 Vs 14:10:14 Integrated Voltage Average and RMS of Measurements X0 = -0.0200000 m, Delta X = 0.0100000 m VT = -0.013477012 +- 0.000001997 Vs 14:10:14 Integrated Field Strength, Average and RMS of Measurements X0 = -0.0200000 m, dx = 0.0100000 m, Nturns = 10.0 VT = -0.013477012 +- 0.000001997 Vs BL = -0.13477012 +- 0.00001997 Tm 14:10:14 Integrated Voltage Measurement... 14:10:32 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = 0.0100000 m VT = -0.006924740 Vs 14:10:47 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = -0.0100000 m VT = 0.006930731 Vs 14:11:03 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = 0.0100000 m VT = -0.006918361 Vs 14:11:18 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = -0.0100000 m VT = 0.006934141 Vs 14:11:33 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = 0.0100000 m VT = -0.006917098 Vs 14:11:48 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = -0.0100000 m VT = 0.006925039 Vs 14:12:03 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = 0.0100000 m VT = -0.006928386 Vs 14:12:18 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = -0.0100000 m VT = 0.006911398 Vs 14:12:18 Integrated Voltage Average and RMS of Measurements X0 = -0.0100000 m, Delta X = 0.0100000 m VT = -0.006923737 +- 0.000003325 Vs 14:12:18 Integrated Field Strength, Average and RMS of Measurements X0 = -0.0100000 m, dx = 0.0100000 m, Nturns = 10.0 VT = -0.006923737 +- 0.000003325 Vs BL = -0.06923737 +- 0.00003325 Tm 14:12:18 Integrated Voltage Measurement... 14:12:36 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = -0.000385278 Vs 14:12:52 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = 0.000372035 Vs 14:13:07 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = -0.000363863 Vs 14:13:22 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = 0.000384683 Vs 14:13:37 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = -0.000377393 Vs 14:13:52 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = 0.000360123 Vs 14:14:07 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = -0.000382163 Vs 14:14:22 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = 0.000359211 Vs 14:14:22 Integrated Voltage Average and RMS of Measurements X0 = 0.0000000 m, Delta X = 0.0100000 m VT = -0.000373094 +- 0.000003772 Vs 14:14:22 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0000000 m, dx = 0.0100000 m, Nturns = 10.0 VT = -0.000373094 +- 0.000003772 Vs BL = -0.00373094 +- 0.00003772 Tm 14:14:22 Integrated Voltage Measurement... 14:14:41 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = 0.0100000 m VT = 0.006158450 Vs 14:14:56 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = -0.0100000 m VT = -0.006191338 Vs 14:15:11 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = 0.0100000 m VT = 0.006163632 Vs 14:15:26 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = -0.0100000 m VT = -0.006180082 Vs 14:15:41 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = 0.0100000 m VT = 0.006164381 Vs 14:15:56 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = -0.0100000 m VT = -0.006194243 Vs 14:16:11 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = 0.0100000 m VT = 0.006162011 Vs 14:16:26 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = -0.0100000 m VT = -0.006189706 Vs 14:16:26 Integrated Voltage Average and RMS of Measurements X0 = 0.0100000 m, Delta X = 0.0100000 m VT = 0.006175480 +- 0.000002660 Vs 14:16:26 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0100000 m, dx = 0.0100000 m, Nturns = 10.0 VT = 0.006175480 +- 0.000002660 Vs BL = 0.06175480 +- 0.00002660 Tm 14:16:26 Integrated Voltage Measurement... 14:16:44 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = 0.0100000 m VT = 0.012699124 Vs 14:16:59 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = -0.0100000 m VT = -0.012754103 Vs 14:17:15 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = 0.0100000 m VT = 0.012693310 Vs 14:17:30 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = -0.0100000 m VT = -0.012759191 Vs 14:17:45 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = 0.0100000 m VT = 0.012681633 Vs 14:18:00 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = -0.0100000 m VT = -0.012759838 Vs 14:18:15 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = 0.0100000 m VT = 0.012694567 Vs 14:18:30 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = -0.0100000 m VT = -0.012732765 Vs 14:18:30 Integrated Voltage Average and RMS of Measurements X0 = 0.0200000 m, Delta X = 0.0100000 m VT = 0.012721816 +- 0.000005250 Vs 14:18:30 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0200000 m, dx = 0.0100000 m, Nturns = 10.0 VT = 0.012721816 +- 0.000005250 Vs BL = 0.12721816 +- 0.00005250 Tm 14:18:35 BL vs X Fit Coefficients: Cn sigCn n (Tm/m^n) ----- --------+-------- 0 -0.00375 0.00002 1 6.54969 0.00120 Compute Fit Residuals: X0 BL sigBL BLfit BL-BLfit (m) (Tm) (Tm) (Tm) (Tm) ---------- --------+-------- -------- -------- -0.020000 -0.13477 0.00002 -0.13475 -0.00002 -0.010000 -0.06924 0.00003 -0.06925 0.00001 0.000000 -0.00373 0.00004 -0.00375 0.00002 0.010000 0.06175 0.00003 0.06174 0.00001 0.020000 0.12722 0.00005 0.12724 -0.00002 14:18:35 Magnet Strength Calculation Harmonic Number = 2 Number Of X0 Positions = 5 X0 BL sigBL (m) (Tm) (Tm) -------- --------+-------- -0.0200 -0.13477 0.00002 -0.0100 -0.06924 0.00003 0.0000 -0.00373 0.00004 0.0100 0.06175 0.00003 0.0200 0.12722 0.00005 SL = 6.5496874 +- 0.0012015 Tm^2/m^n 14:18:35 Quadrupole Strength Calculation Check Slope = 6.549687 Tm/m, Inter = -0.003753 Tm, Sig_slope = 0.000779 Tm/m, Est_sig = 0.000025 Tm GL = 6.549687 +- 0.000779 T 14:18:35 Magnet Strength Measurement Harmonic Number = 2 SL = 6.5496874 +- 0.0012015 Tm^2/m^n 14:18:36 Magnet Current, Imag = 50.2698 A Vtrans = 0.502698 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 14:18:38 Magnet Current, Imag = 50.2698 A Vtrans = 0.502698 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 14:18:39 Magnet Current, Imag = 50.2693 A Vtrans = 0.502693 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 14:18:40 Magnet Current, Imag = 50.2695 A Vtrans = 0.502695 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 14:18:40 Magnet Current Measurement: Imag = 50.2696 +- 0.0002 A 14:18:40 Magnet current ramping to 55.0000 A... 14:18:49 Magnet Current, Imag = 55.3006 A Vtrans = 0.553006 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 14:18:50 Magnet Current, Imag = 55.2998 A Vtrans = 0.552998 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 14:18:52 Magnet Current, Imag = 55.2979 A Vtrans = 0.552979 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 14:18:53 Magnet Current, Imag = 55.2986 A Vtrans = 0.552986 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 14:18:53 Magnet Current Measurement: Imag = 55.2992 +- 0.0010 A 14:18:53 Magnetic Field Measurement Measuring the magnet strength... 14:18:53 Integrated Voltage Measurement... 14:19:15 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = 0.0100000 m VT = -0.014847607 Vs 14:19:30 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = -0.0100000 m VT = 0.014802801 Vs 14:19:45 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = 0.0100000 m VT = -0.014849452 Vs 14:20:00 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = -0.0100000 m VT = 0.014783258 Vs 14:20:15 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = 0.0100000 m VT = -0.014862667 Vs 14:20:30 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = -0.0100000 m VT = 0.014787537 Vs 14:20:46 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = 0.0100000 m VT = -0.014849118 Vs 14:21:01 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = -0.0100000 m VT = 0.014795329 Vs 14:21:01 Integrated Voltage Average and RMS of Measurements X0 = -0.0200000 m, Delta X = 0.0100000 m VT = -0.014822221 +- 0.000003592 Vs 14:21:01 Integrated Field Strength, Average and RMS of Measurements X0 = -0.0200000 m, dx = 0.0100000 m, Nturns = 10.0 VT = -0.014822221 +- 0.000003592 Vs BL = -0.14822221 +- 0.00003592 Tm 14:21:01 Integrated Voltage Measurement... 14:21:19 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = 0.0100000 m VT = -0.007637793 Vs 14:21:34 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = -0.0100000 m VT = 0.007601868 Vs 14:21:49 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = 0.0100000 m VT = -0.007617064 Vs 14:22:04 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = -0.0100000 m VT = 0.007607431 Vs 14:22:19 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = 0.0100000 m VT = -0.007622471 Vs 14:22:34 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = -0.0100000 m VT = 0.007621726 Vs 14:22:49 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = 0.0100000 m VT = -0.007621392 Vs 14:23:04 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = -0.0100000 m VT = 0.007604774 Vs 14:23:04 Integrated Voltage Average and RMS of Measurements X0 = -0.0100000 m, Delta X = 0.0100000 m VT = -0.007616815 +- 0.000004237 Vs 14:23:05 Integrated Field Strength, Average and RMS of Measurements X0 = -0.0100000 m, dx = 0.0100000 m, Nturns = 10.0 VT = -0.007616815 +- 0.000004237 Vs BL = -0.07616815 +- 0.00004237 Tm 14:23:05 Integrated Voltage Measurement... 14:23:23 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = -0.000436069 Vs 14:23:38 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = 0.000384680 Vs 14:23:53 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = -0.000426503 Vs 14:24:08 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = 0.000400293 Vs 14:24:23 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = -0.000426559 Vs 14:24:38 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = 0.000395726 Vs 14:24:54 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = -0.000417938 Vs 14:25:09 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = 0.000406494 Vs 14:25:09 Integrated Voltage Average and RMS of Measurements X0 = 0.0000000 m, Delta X = 0.0100000 m VT = -0.000411783 +- 0.000001139 Vs 14:25:09 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0000000 m, dx = 0.0100000 m, Nturns = 10.0 VT = -0.000411783 +- 0.000001139 Vs BL = -0.00411783 +- 0.00001139 Tm 14:25:09 Integrated Voltage Measurement... 14:25:28 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = 0.0100000 m VT = 0.006794150 Vs 14:25:43 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = -0.0100000 m VT = -0.006783080 Vs 14:25:58 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = 0.0100000 m VT = 0.006793178 Vs 14:26:13 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = -0.0100000 m VT = -0.006801936 Vs 14:26:28 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = 0.0100000 m VT = 0.006785554 Vs 14:26:43 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = -0.0100000 m VT = -0.006797344 Vs 14:26:58 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = 0.0100000 m VT = 0.006777898 Vs 14:27:13 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = -0.0100000 m VT = -0.006803298 Vs 14:27:13 Integrated Voltage Average and RMS of Measurements X0 = 0.0100000 m, Delta X = 0.0100000 m VT = 0.006792055 +- 0.000003339 Vs 14:27:13 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0100000 m, dx = 0.0100000 m, Nturns = 10.0 VT = 0.006792055 +- 0.000003339 Vs BL = 0.06792055 +- 0.00003339 Tm 14:27:13 Integrated Voltage Measurement... 14:27:32 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = 0.0100000 m VT = 0.013990508 Vs 14:27:47 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = -0.0100000 m VT = -0.013983846 Vs 14:28:02 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = 0.0100000 m VT = 0.014000754 Vs 14:28:17 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = -0.0100000 m VT = -0.013979165 Vs 14:28:32 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = 0.0100000 m VT = 0.014005304 Vs 14:28:47 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = -0.0100000 m VT = -0.013978836 Vs 14:29:02 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = 0.0100000 m VT = 0.013999440 Vs 14:29:17 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = -0.0100000 m VT = -0.013983689 Vs 14:29:17 Integrated Voltage Average and RMS of Measurements X0 = 0.0200000 m, Delta X = 0.0100000 m VT = 0.013990193 +- 0.000001908 Vs 14:29:17 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0200000 m, dx = 0.0100000 m, Nturns = 10.0 VT = 0.013990193 +- 0.000001908 Vs BL = 0.13990193 +- 0.00001908 Tm 14:29:23 BL vs X Fit Coefficients: Cn sigCn n (Tm/m^n) ----- --------+-------- 0 -0.00414 0.00001 1 7.20337 0.00098 Compute Fit Residuals: X0 BL sigBL BLfit BL-BLfit (m) (Tm) (Tm) (Tm) (Tm) ---------- --------+-------- -------- -------- -0.020000 -0.14822 0.00004 -0.14820 -0.00002 -0.010000 -0.07617 0.00004 -0.07617 0.00000 0.000000 -0.00412 0.00001 -0.00414 0.00002 0.010000 0.06792 0.00003 0.06790 0.00002 0.020000 0.13990 0.00002 0.13993 -0.00003 14:29:23 Magnet Strength Calculation Harmonic Number = 2 Number Of X0 Positions = 5 X0 BL sigBL (m) (Tm) (Tm) -------- --------+-------- -0.0200 -0.14822 0.00004 -0.0100 -0.07617 0.00004 0.0000 -0.00412 0.00001 0.0100 0.06792 0.00003 0.0200 0.13990 0.00002 SL = 7.2033697 +- 0.0009760 Tm^2/m^n 14:29:23 Quadrupole Strength Calculation Check Slope = 7.203370 Tm/m, Inter = -0.004137 Tm, Sig_slope = 0.000831 Tm/m, Est_sig = 0.000026 Tm GL = 7.203370 +- 0.000831 T 14:29:23 Magnet Strength Measurement Harmonic Number = 2 SL = 7.2033697 +- 0.0009760 Tm^2/m^n 14:29:24 Magnet Current, Imag = 55.2999 A Vtrans = 0.552999 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 14:29:25 Magnet Current, Imag = 55.3001 A Vtrans = 0.553001 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 14:29:27 Magnet Current, Imag = 55.2998 A Vtrans = 0.552998 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 14:29:28 Magnet Current, Imag = 55.2989 A Vtrans = 0.552989 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 14:29:28 Magnet Current Measurement: Imag = 55.2997 +- 0.0005 A 14:29:28 Magnet current ramping to 60.0000 A... 14:29:37 Magnet Current, Imag = 60.2881 A Vtrans = 0.602881 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 14:29:38 Magnet Current, Imag = 60.2880 A Vtrans = 0.602880 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 14:29:39 Magnet Current, Imag = 60.2861 A Vtrans = 0.602861 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 14:29:40 Magnet Current, Imag = 60.2870 A Vtrans = 0.602870 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 14:29:40 Magnet Current Measurement: Imag = 60.2873 +- 0.0008 A 14:29:41 Magnetic Field Measurement Measuring the magnet strength... 14:29:41 Integrated Voltage Measurement... 14:30:03 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = 0.0100000 m VT = -0.016141366 Vs 14:30:18 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = -0.0100000 m VT = 0.016148524 Vs 14:30:33 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = 0.0100000 m VT = -0.016154687 Vs 14:30:48 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = -0.0100000 m VT = 0.016132472 Vs 14:31:04 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = 0.0100000 m VT = -0.016187674 Vs 14:31:19 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = -0.0100000 m VT = 0.016129041 Vs 14:31:34 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = 0.0100000 m VT = -0.016163935 Vs 14:31:49 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = -0.0100000 m VT = 0.016143458 Vs 14:31:49 Integrated Voltage Average and RMS of Measurements X0 = -0.0200000 m, Delta X = 0.0100000 m VT = -0.016150145 +- 0.000006128 Vs 14:31:49 Integrated Field Strength, Average and RMS of Measurements X0 = -0.0200000 m, dx = 0.0100000 m, Nturns = 10.0 VT = -0.016150145 +- 0.000006128 Vs BL = -0.16150145 +- 0.00006128 Tm 14:31:49 Integrated Voltage Measurement... 14:32:08 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = 0.0100000 m VT = -0.008305486 Vs 14:32:23 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = -0.0100000 m VT = 0.008281063 Vs 14:32:38 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = 0.0100000 m VT = -0.008309460 Vs 14:32:54 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = -0.0100000 m VT = 0.008290951 Vs 14:33:09 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = 0.0100000 m VT = -0.008295765 Vs 14:33:24 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = -0.0100000 m VT = 0.008297896 Vs 14:33:39 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = 0.0100000 m VT = -0.008306179 Vs 14:33:54 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = -0.0100000 m VT = 0.008280646 Vs 14:33:54 Integrated Voltage Average and RMS of Measurements X0 = -0.0100000 m, Delta X = 0.0100000 m VT = -0.008295931 +- 0.000002850 Vs 14:33:54 Integrated Field Strength, Average and RMS of Measurements X0 = -0.0100000 m, dx = 0.0100000 m, Nturns = 10.0 VT = -0.008295931 +- 0.000002850 Vs BL = -0.08295931 +- 0.00002850 Tm 14:33:54 Integrated Voltage Measurement... 14:34:12 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = -0.000449495 Vs 14:34:27 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = 0.000445279 Vs 14:34:42 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = -0.000447260 Vs 14:34:57 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = 0.000439538 Vs 14:35:12 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = -0.000453166 Vs 14:35:28 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = 0.000435827 Vs 14:35:43 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = -0.000448145 Vs 14:35:58 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = 0.000458663 Vs 14:35:58 Integrated Voltage Average and RMS of Measurements X0 = 0.0000000 m, Delta X = 0.0100000 m VT = -0.000447172 +- 0.000003882 Vs 14:35:58 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0000000 m, dx = 0.0100000 m, Nturns = 10.0 VT = -0.000447172 +- 0.000003882 Vs BL = -0.00447172 +- 0.00003882 Tm 14:35:58 Integrated Voltage Measurement... 14:36:16 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = 0.0100000 m VT = 0.007394384 Vs 14:36:31 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = -0.0100000 m VT = -0.007401320 Vs 14:36:46 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = 0.0100000 m VT = 0.007402111 Vs 14:37:01 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = -0.0100000 m VT = -0.007407396 Vs 14:37:16 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = 0.0100000 m VT = 0.007369854 Vs 14:37:32 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = -0.0100000 m VT = -0.007429783 Vs 14:37:47 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = 0.0100000 m VT = 0.007375930 Vs 14:38:02 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = -0.0100000 m VT = -0.007416866 Vs 14:38:02 Integrated Voltage Average and RMS of Measurements X0 = 0.0100000 m, Delta X = 0.0100000 m VT = 0.007399706 +- 0.000003157 Vs 14:38:02 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0100000 m, dx = 0.0100000 m, Nturns = 10.0 VT = 0.007399706 +- 0.000003157 Vs BL = 0.07399706 +- 0.00003157 Tm 14:38:02 Integrated Voltage Measurement... 14:38:20 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = 0.0100000 m VT = 0.015258666 Vs 14:38:35 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = -0.0100000 m VT = -0.015241763 Vs 14:38:51 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = 0.0100000 m VT = 0.015243078 Vs 14:39:06 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = -0.0100000 m VT = -0.015265668 Vs 14:39:21 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = 0.0100000 m VT = 0.015243067 Vs 14:39:36 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = -0.0100000 m VT = -0.015236743 Vs 14:39:51 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = 0.0100000 m VT = 0.015254675 Vs 14:40:06 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = -0.0100000 m VT = -0.015252027 Vs 14:40:06 Integrated Voltage Average and RMS of Measurements X0 = 0.0200000 m, Delta X = 0.0100000 m VT = 0.015249461 +- 0.000005726 Vs 14:40:06 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0200000 m, dx = 0.0100000 m, Nturns = 10.0 VT = 0.015249461 +- 0.000005726 Vs BL = 0.15249461 +- 0.00005726 Tm 14:40:11 BL vs X Fit Coefficients: Cn sigCn n (Tm/m^n) ----- --------+-------- 0 -0.00449 0.00002 1 7.84948 0.00173 Compute Fit Residuals: X0 BL sigBL BLfit BL-BLfit (m) (Tm) (Tm) (Tm) (Tm) ---------- --------+-------- -------- -------- -0.020000 -0.16150 0.00006 -0.16148 -0.00002 -0.010000 -0.08296 0.00003 -0.08298 0.00002 0.000000 -0.00447 0.00004 -0.00449 0.00002 0.010000 0.07400 0.00003 0.07401 -0.00001 0.020000 0.15249 0.00006 0.15250 -0.00001 14:40:11 Magnet Strength Calculation Harmonic Number = 2 Number Of X0 Positions = 5 X0 BL sigBL (m) (Tm) (Tm) -------- --------+-------- -0.0200 -0.16150 0.00006 -0.0100 -0.08296 0.00003 0.0000 -0.00447 0.00004 0.0100 0.07400 0.00003 0.0200 0.15249 0.00006 SL = 7.8494848 +- 0.0017304 Tm^2/m^n 14:40:11 Quadrupole Strength Calculation Check Slope = 7.849485 Tm/m, Inter = -0.004488 Tm, Sig_slope = 0.000714 Tm/m, Est_sig = 0.000023 Tm GL = 7.849485 +- 0.000714 T 14:40:11 Magnet Strength Measurement Harmonic Number = 2 SL = 7.8494848 +- 0.0017304 Tm^2/m^n 14:40:13 Magnet Current, Imag = 60.2884 A Vtrans = 0.602884 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 14:40:14 Magnet Current, Imag = 60.2887 A Vtrans = 0.602887 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 14:40:15 Magnet Current, Imag = 60.2877 A Vtrans = 0.602877 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 14:40:17 Magnet Current, Imag = 60.2879 A Vtrans = 0.602879 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 14:40:17 Magnet Current Measurement: Imag = 60.2882 +- 0.0004 A 14:40:17 Magnet current ramping to 65.0000 A... 14:40:25 Magnet Current, Imag = 65.3408 A Vtrans = 0.653408 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 14:40:27 Magnet Current, Imag = 65.3416 A Vtrans = 0.653416 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 14:40:28 Magnet Current, Imag = 65.3395 A Vtrans = 0.653395 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 14:40:29 Magnet Current, Imag = 65.3390 A Vtrans = 0.653390 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 14:40:29 Magnet Current Measurement: Imag = 65.3402 +- 0.0010 A 14:40:29 Magnetic Field Measurement Measuring the magnet strength... 14:40:29 Integrated Voltage Measurement... 14:40:52 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = 0.0100000 m VT = -0.017495854 Vs 14:41:07 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = -0.0100000 m VT = 0.017486618 Vs 14:41:22 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = 0.0100000 m VT = -0.017514382 Vs 14:41:37 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = -0.0100000 m VT = 0.017463745 Vs 14:41:52 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = 0.0100000 m VT = -0.017511154 Vs 14:42:08 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = -0.0100000 m VT = 0.017471324 Vs 14:42:23 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = 0.0100000 m VT = -0.017514879 Vs 14:42:38 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = -0.0100000 m VT = 0.017463979 Vs 14:42:38 Integrated Voltage Average and RMS of Measurements X0 = -0.0200000 m, Delta X = 0.0100000 m VT = -0.017490242 +- 0.000001004 Vs 14:42:38 Integrated Field Strength, Average and RMS of Measurements X0 = -0.0200000 m, dx = 0.0100000 m, Nturns = 10.0 VT = -0.017490242 +- 0.000001004 Vs BL = -0.17490242 +- 0.00001004 Tm 14:42:38 Integrated Voltage Measurement... 14:42:57 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = 0.0100000 m VT = -0.009016114 Vs 14:43:12 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = -0.0100000 m VT = 0.008967253 Vs 14:43:27 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = 0.0100000 m VT = -0.008999713 Vs 14:43:42 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = -0.0100000 m VT = 0.008973896 Vs 14:43:57 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = 0.0100000 m VT = -0.008974938 Vs 14:44:13 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = -0.0100000 m VT = 0.009008629 Vs 14:44:28 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = 0.0100000 m VT = -0.008964870 Vs 14:44:43 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = -0.0100000 m VT = 0.008999878 Vs 14:44:43 Integrated Voltage Average and RMS of Measurements X0 = -0.0100000 m, Delta X = 0.0100000 m VT = -0.008988161 +- 0.000003901 Vs 14:44:43 Integrated Field Strength, Average and RMS of Measurements X0 = -0.0100000 m, dx = 0.0100000 m, Nturns = 10.0 VT = -0.008988161 +- 0.000003901 Vs BL = -0.08988161 +- 0.00003901 Tm 14:44:43 Integrated Voltage Measurement... 14:45:02 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = -0.000458507 Vs 14:45:17 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = 0.000508394 Vs 14:45:32 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = -0.000451132 Vs 14:45:47 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = 0.000509229 Vs 14:46:02 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = -0.000460866 Vs 14:46:17 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = 0.000495074 Vs 14:46:32 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = -0.000470973 Vs 14:46:47 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = 0.000496660 Vs 14:46:48 Integrated Voltage Average and RMS of Measurements X0 = 0.0000000 m, Delta X = 0.0100000 m VT = -0.000481354 +- 0.000002413 Vs 14:46:48 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0000000 m, dx = 0.0100000 m, Nturns = 10.0 VT = -0.000481354 +- 0.000002413 Vs BL = -0.00481354 +- 0.00002413 Tm 14:46:48 Integrated Voltage Measurement... 14:47:06 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = 0.0100000 m VT = 0.008023435 Vs 14:47:21 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = -0.0100000 m VT = -0.008018729 Vs 14:47:36 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = 0.0100000 m VT = 0.008032739 Vs 14:47:51 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = -0.0100000 m VT = -0.007991891 Vs 14:48:07 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = 0.0100000 m VT = 0.008039807 Vs 14:48:22 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = -0.0100000 m VT = -0.008007415 Vs 14:48:37 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = 0.0100000 m VT = 0.008023935 Vs 14:48:52 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = -0.0100000 m VT = -0.008006847 Vs 14:48:52 Integrated Voltage Average and RMS of Measurements X0 = 0.0100000 m, Delta X = 0.0100000 m VT = 0.008018100 +- 0.000004474 Vs 14:48:52 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0100000 m, dx = 0.0100000 m, Nturns = 10.0 VT = 0.008018100 +- 0.000004474 Vs BL = 0.08018100 +- 0.00004474 Tm 14:48:52 Integrated Voltage Measurement... 14:49:10 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = 0.0100000 m VT = 0.016518688 Vs 14:49:25 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = -0.0100000 m VT = -0.016517925 Vs 14:49:40 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = 0.0100000 m VT = 0.016514948 Vs 14:49:55 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = -0.0100000 m VT = -0.016522059 Vs 14:50:10 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = 0.0100000 m VT = 0.016531742 Vs 14:50:26 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = -0.0100000 m VT = -0.016491127 Vs 14:50:41 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = 0.0100000 m VT = 0.016524437 Vs 14:50:56 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = -0.0100000 m VT = -0.016505674 Vs 14:50:56 Integrated Voltage Average and RMS of Measurements X0 = 0.0200000 m, Delta X = 0.0100000 m VT = 0.016515825 +- 0.000002881 Vs 14:50:56 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0200000 m, dx = 0.0100000 m, Nturns = 10.0 VT = 0.016515825 +- 0.000002881 Vs BL = 0.16515825 +- 0.00002881 Tm 14:51:01 BL vs X Fit Coefficients: Cn sigCn n (Tm/m^n) ----- --------+-------- 0 -0.00485 0.00001 1 8.50184 0.00085 Compute Fit Residuals: X0 BL sigBL BLfit BL-BLfit (m) (Tm) (Tm) (Tm) (Tm) ---------- --------+-------- -------- -------- -0.020000 -0.17490 0.00001 -0.17489 -0.00001 -0.010000 -0.08988 0.00004 -0.08987 -0.00001 0.000000 -0.00481 0.00002 -0.00485 0.00004 0.010000 0.08018 0.00004 0.08017 0.00001 0.020000 0.16516 0.00003 0.16519 -0.00003 14:51:01 Magnet Strength Calculation Harmonic Number = 2 Number Of X0 Positions = 5 X0 BL sigBL (m) (Tm) (Tm) -------- --------+-------- -0.0200 -0.17490 0.00001 -0.0100 -0.08988 0.00004 0.0000 -0.00481 0.00002 0.0100 0.08018 0.00004 0.0200 0.16516 0.00003 SL = 8.5018395 +- 0.0008513 Tm^2/m^n 14:51:01 Quadrupole Strength Calculation Check Slope = 8.501839 Tm/m, Inter = -0.004852 Tm, Sig_slope = 0.000950 Tm/m, Est_sig = 0.000030 Tm GL = 8.501839 +- 0.000950 T 14:51:01 Magnet Strength Measurement Harmonic Number = 2 SL = 8.5018395 +- 0.0008513 Tm^2/m^n 14:51:02 Magnet Current, Imag = 65.3417 A Vtrans = 0.653417 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 14:51:04 Magnet Current, Imag = 65.3411 A Vtrans = 0.653411 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 14:51:05 Magnet Current, Imag = 65.3413 A Vtrans = 0.653413 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 14:51:06 Magnet Current, Imag = 65.3410 A Vtrans = 0.653410 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 14:51:06 Magnet Current Measurement: Imag = 65.3413 +- 0.0003 A 14:51:06 Magnet current ramping to 70.0000 A... 14:51:15 Magnet Current, Imag = 70.3648 A Vtrans = 0.703648 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 14:51:16 Magnet Current, Imag = 70.3641 A Vtrans = 0.703641 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 14:51:18 Magnet Current, Imag = 70.3636 A Vtrans = 0.703636 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 14:51:19 Magnet Current, Imag = 70.3630 A Vtrans = 0.703630 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 14:51:19 Magnet Current Measurement: Imag = 70.3639 +- 0.0007 A 14:51:19 Magnetic Field Measurement Measuring the magnet strength... 14:51:19 Integrated Voltage Measurement... 14:51:41 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = 0.0100000 m VT = -0.018807599 Vs 14:51:57 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = -0.0100000 m VT = 0.018821117 Vs 14:52:12 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = 0.0100000 m VT = -0.018819872 Vs 14:52:27 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = -0.0100000 m VT = 0.018819684 Vs 14:52:42 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = 0.0100000 m VT = -0.018820049 Vs 14:52:57 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = -0.0100000 m VT = 0.018828873 Vs 14:53:12 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = 0.0100000 m VT = -0.018809755 Vs 14:53:27 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = -0.0100000 m VT = 0.018818831 Vs 14:53:27 Integrated Voltage Average and RMS of Measurements X0 = -0.0200000 m, Delta X = 0.0100000 m VT = -0.018818222 +- 0.000004234 Vs 14:53:28 Integrated Field Strength, Average and RMS of Measurements X0 = -0.0200000 m, dx = 0.0100000 m, Nturns = 10.0 VT = -0.018818222 +- 0.000004234 Vs BL = -0.18818222 +- 0.00004234 Tm 14:53:28 Integrated Voltage Measurement... 14:53:46 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = 0.0100000 m VT = -0.009674401 Vs 14:54:02 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = -0.0100000 m VT = 0.009673788 Vs 14:54:17 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = 0.0100000 m VT = -0.009656475 Vs 14:54:32 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = -0.0100000 m VT = 0.009684753 Vs 14:54:47 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = 0.0100000 m VT = -0.009635323 Vs 14:55:02 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = -0.0100000 m VT = 0.009714535 Vs 14:55:17 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = 0.0100000 m VT = -0.009623699 Vs 14:55:32 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = -0.0100000 m VT = 0.009711359 Vs 14:55:32 Integrated Voltage Average and RMS of Measurements X0 = -0.0100000 m, Delta X = 0.0100000 m VT = -0.009671792 +- 0.000002945 Vs 14:55:32 Integrated Field Strength, Average and RMS of Measurements X0 = -0.0100000 m, dx = 0.0100000 m, Nturns = 10.0 VT = -0.009671792 +- 0.000002945 Vs BL = -0.09671792 +- 0.00002945 Tm 14:55:32 Integrated Voltage Measurement... 14:55:51 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = -0.000483138 Vs 14:56:06 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = 0.000555299 Vs 14:56:21 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = -0.000482638 Vs 14:56:36 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = 0.000550287 Vs 14:56:51 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = -0.000489079 Vs 14:57:06 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = 0.000547298 Vs 14:57:21 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = -0.000495447 Vs 14:57:36 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = 0.000545065 Vs 14:57:36 Integrated Voltage Average and RMS of Measurements X0 = 0.0000000 m, Delta X = 0.0100000 m VT = -0.000518531 +- 0.000001400 Vs 14:57:36 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0000000 m, dx = 0.0100000 m, Nturns = 10.0 VT = -0.000518531 +- 0.000001400 Vs BL = -0.00518531 +- 0.00001400 Tm 14:57:36 Integrated Voltage Measurement... 14:57:55 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = 0.0100000 m VT = 0.008649615 Vs 14:58:10 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = -0.0100000 m VT = -0.008598965 Vs 14:58:25 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = 0.0100000 m VT = 0.008656821 Vs 14:58:40 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = -0.0100000 m VT = -0.008593437 Vs 14:58:56 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = 0.0100000 m VT = 0.008651037 Vs 14:59:11 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = -0.0100000 m VT = -0.008600568 Vs 14:59:26 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = 0.0100000 m VT = 0.008645010 Vs 14:59:41 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = -0.0100000 m VT = -0.008593423 Vs 14:59:41 Integrated Voltage Average and RMS of Measurements X0 = 0.0100000 m, Delta X = 0.0100000 m VT = 0.008623609 +- 0.000002592 Vs 14:59:41 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0100000 m, dx = 0.0100000 m, Nturns = 10.0 VT = 0.008623609 +- 0.000002592 Vs BL = 0.08623609 +- 0.00002592 Tm 14:59:41 Integrated Voltage Measurement... 14:59:59 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = 0.0100000 m VT = 0.017786531 Vs 15:00:14 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = -0.0100000 m VT = -0.017740592 Vs 15:00:29 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = 0.0100000 m VT = 0.017779615 Vs 15:00:45 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = -0.0100000 m VT = -0.017744651 Vs 15:01:00 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = 0.0100000 m VT = 0.017794677 Vs 15:01:15 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = -0.0100000 m VT = -0.017737651 Vs 15:01:30 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = 0.0100000 m VT = 0.017802991 Vs 15:01:45 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = -0.0100000 m VT = -0.017726755 Vs 15:01:45 Integrated Voltage Average and RMS of Measurements X0 = 0.0200000 m, Delta X = 0.0100000 m VT = 0.017764183 +- 0.000001499 Vs 15:01:45 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0200000 m, dx = 0.0100000 m, Nturns = 10.0 VT = 0.017764183 +- 0.000001499 Vs BL = 0.17764183 +- 0.00001499 Tm 15:01:51 BL vs X Fit Coefficients: Cn sigCn n (Tm/m^n) ----- --------+-------- 0 -0.00524 0.00001 1 9.14602 0.00098 Compute Fit Residuals: X0 BL sigBL BLfit BL-BLfit (m) (Tm) (Tm) (Tm) (Tm) ---------- --------+-------- -------- -------- -0.020000 -0.18818 0.00004 -0.18816 -0.00002 -0.010000 -0.09672 0.00003 -0.09670 -0.00002 0.000000 -0.00519 0.00001 -0.00524 0.00006 0.010000 0.08624 0.00003 0.08622 0.00002 0.020000 0.17764 0.00001 0.17768 -0.00004 15:01:51 Magnet Strength Calculation Harmonic Number = 2 Number Of X0 Positions = 5 X0 BL sigBL (m) (Tm) (Tm) -------- --------+-------- -0.0200 -0.18818 0.00004 -0.0100 -0.09672 0.00003 0.0000 -0.00519 0.00001 0.0100 0.08624 0.00003 0.0200 0.17764 0.00001 SL = 9.1460212 +- 0.0009803 Tm^2/m^n 15:01:51 Quadrupole Strength Calculation Check Slope = 9.146021 Tm/m, Inter = -0.005242 Tm, Sig_slope = 0.001355 Tm/m, Est_sig = 0.000043 Tm GL = 9.146021 +- 0.001355 T 15:01:51 Magnet Strength Measurement Harmonic Number = 2 SL = 9.1460212 +- 0.0009803 Tm^2/m^n 15:01:52 Magnet Current, Imag = 70.3651 A Vtrans = 0.703651 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 15:01:53 Magnet Current, Imag = 70.3659 A Vtrans = 0.703659 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 15:01:55 Magnet Current, Imag = 70.3654 A Vtrans = 0.703654 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 15:01:56 Magnet Current, Imag = 70.3660 A Vtrans = 0.703660 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 15:01:56 Magnet Current Measurement: Imag = 70.3656 +- 0.0004 A 15:01:56 Magnet current ramping to 75.0000 A... 15:02:05 Magnet Current, Imag = 75.4451 A Vtrans = 0.754451 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 15:02:06 Magnet Current, Imag = 75.4455 A Vtrans = 0.754455 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 15:02:07 Magnet Current, Imag = 75.4438 A Vtrans = 0.754438 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 15:02:09 Magnet Current, Imag = 75.4437 A Vtrans = 0.754437 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 15:02:09 Magnet Current Measurement: Imag = 75.4445 +- 0.0008 A 15:02:09 Magnetic Field Measurement Measuring the magnet strength... 15:02:09 Integrated Voltage Measurement... 15:02:31 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = 0.0100000 m VT = -0.020087432 Vs 15:02:46 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = -0.0100000 m VT = 0.020186498 Vs 15:03:01 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = 0.0100000 m VT = -0.020089499 Vs 15:03:16 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = -0.0100000 m VT = 0.020178893 Vs 15:03:31 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = 0.0100000 m VT = -0.020097148 Vs 15:03:46 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = -0.0100000 m VT = 0.020177855 Vs 15:04:01 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = 0.0100000 m VT = -0.020108539 Vs 15:04:17 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = -0.0100000 m VT = 0.020170135 Vs 15:04:17 Integrated Voltage Average and RMS of Measurements X0 = -0.0200000 m, Delta X = 0.0100000 m VT = -0.020137000 +- 0.000001842 Vs 15:04:17 Integrated Field Strength, Average and RMS of Measurements X0 = -0.0200000 m, dx = 0.0100000 m, Nturns = 10.0 VT = -0.020137000 +- 0.000001842 Vs BL = -0.20137000 +- 0.00001842 Tm 15:04:17 Integrated Voltage Measurement... 15:04:35 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = 0.0100000 m VT = -0.010321963 Vs 15:04:50 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = -0.0100000 m VT = 0.010363377 Vs 15:05:05 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = 0.0100000 m VT = -0.010322808 Vs 15:05:20 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = -0.0100000 m VT = 0.010375261 Vs 15:05:35 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = 0.0100000 m VT = -0.010327948 Vs 15:05:50 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = -0.0100000 m VT = 0.010367128 Vs 15:06:05 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = 0.0100000 m VT = -0.010321710 Vs 15:06:20 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = -0.0100000 m VT = 0.010364672 Vs 15:06:20 Integrated Voltage Average and RMS of Measurements X0 = -0.0100000 m, Delta X = 0.0100000 m VT = -0.010345608 +- 0.000002736 Vs 15:06:20 Integrated Field Strength, Average and RMS of Measurements X0 = -0.0100000 m, dx = 0.0100000 m, Nturns = 10.0 VT = -0.010345608 +- 0.000002736 Vs BL = -0.10345608 +- 0.00002736 Tm 15:06:21 Integrated Voltage Measurement... 15:06:39 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = -0.000545706 Vs 15:06:54 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = 0.000576125 Vs 15:07:09 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = -0.000534455 Vs 15:07:24 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = 0.000575564 Vs 15:07:39 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = -0.000535657 Vs 15:07:54 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = 0.000568559 Vs 15:08:09 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = -0.000545276 Vs 15:08:24 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = 0.000567780 Vs 15:08:24 Integrated Voltage Average and RMS of Measurements X0 = 0.0000000 m, Delta X = 0.0100000 m VT = -0.000556140 +- 0.000003182 Vs 15:08:24 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0000000 m, dx = 0.0100000 m, Nturns = 10.0 VT = -0.000556140 +- 0.000003182 Vs BL = -0.00556140 +- 0.00003182 Tm 15:08:25 Integrated Voltage Measurement... 15:08:43 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = 0.0100000 m VT = 0.009244462 Vs 15:08:58 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = -0.0100000 m VT = -0.009225038 Vs 15:09:13 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = 0.0100000 m VT = 0.009241094 Vs 15:09:28 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = -0.0100000 m VT = -0.009219941 Vs 15:09:43 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = 0.0100000 m VT = 0.009230217 Vs 15:09:58 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = -0.0100000 m VT = -0.009227497 Vs 15:10:13 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = 0.0100000 m VT = 0.009230834 Vs 15:10:28 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = -0.0100000 m VT = -0.009235771 Vs 15:10:28 Integrated Voltage Average and RMS of Measurements X0 = 0.0100000 m, Delta X = 0.0100000 m VT = 0.009231857 +- 0.000002305 Vs 15:10:28 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0100000 m, dx = 0.0100000 m, Nturns = 10.0 VT = 0.009231857 +- 0.000002305 Vs BL = 0.09231857 +- 0.00002305 Tm 15:10:28 Integrated Voltage Measurement... 15:10:47 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = 0.0100000 m VT = 0.019006054 Vs 15:11:02 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = -0.0100000 m VT = -0.019020840 Vs 15:11:17 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = 0.0100000 m VT = 0.019002362 Vs 15:11:32 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = -0.0100000 m VT = -0.019019609 Vs 15:11:47 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = 0.0100000 m VT = 0.019003771 Vs 15:12:02 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = -0.0100000 m VT = -0.019014550 Vs 15:12:17 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = 0.0100000 m VT = 0.019015301 Vs 15:12:32 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = -0.0100000 m VT = -0.019005226 Vs 15:12:32 Integrated Voltage Average and RMS of Measurements X0 = 0.0200000 m, Delta X = 0.0100000 m VT = 0.019010964 +- 0.000001574 Vs 15:12:32 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0200000 m, dx = 0.0100000 m, Nturns = 10.0 VT = 0.019010964 +- 0.000001574 Vs BL = 0.19010964 +- 0.00001574 Tm 15:12:38 BL vs X Fit Coefficients: Cn sigCn n (Tm/m^n) ----- --------+-------- 0 -0.00559 0.00001 1 9.78734 0.00060 Compute Fit Residuals: X0 BL sigBL BLfit BL-BLfit (m) (Tm) (Tm) (Tm) (Tm) ---------- --------+-------- -------- -------- -0.020000 -0.20137 0.00002 -0.20134 -0.00003 -0.010000 -0.10346 0.00003 -0.10347 0.00001 0.000000 -0.00556 0.00003 -0.00559 0.00003 0.010000 0.09232 0.00002 0.09228 0.00004 0.020000 0.19011 0.00002 0.19015 -0.00005 15:12:38 Magnet Strength Calculation Harmonic Number = 2 Number Of X0 Positions = 5 X0 BL sigBL (m) (Tm) (Tm) -------- --------+-------- -0.0200 -0.20137 0.00002 -0.0100 -0.10346 0.00003 0.0000 -0.00556 0.00003 0.0100 0.09232 0.00002 0.0200 0.19011 0.00002 SL = 9.7873393 +- 0.0006024 Tm^2/m^n 15:12:38 Quadrupole Strength Calculation Check Slope = 9.787339 Tm/m, Inter = -0.005592 Tm, Sig_slope = 0.001344 Tm/m, Est_sig = 0.000042 Tm GL = 9.787339 +- 0.001344 T 15:12:38 Magnet Strength Measurement Harmonic Number = 2 SL = 9.7873393 +- 0.0006024 Tm^2/m^n 15:12:39 Magnet Current, Imag = 75.4451 A Vtrans = 0.754451 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 15:12:40 Magnet Current, Imag = 75.4449 A Vtrans = 0.754449 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 15:12:42 Magnet Current, Imag = 75.4445 A Vtrans = 0.754445 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 15:12:43 Magnet Current, Imag = 75.4446 A Vtrans = 0.754446 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 15:12:43 Magnet Current Measurement: Imag = 75.4448 +- 0.0002 A 15:12:43 Magnet current ramping to 80.0000 A... 15:12:52 Magnet Current, Imag = 80.4972 A Vtrans = 0.804972 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 15:12:53 Magnet Current, Imag = 80.4986 A Vtrans = 0.804986 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 15:12:54 Magnet Current, Imag = 80.4972 A Vtrans = 0.804972 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 15:12:56 Magnet Current, Imag = 80.4980 A Vtrans = 0.804980 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 15:12:56 Magnet Current Measurement: Imag = 80.4977 +- 0.0006 A 15:12:56 Magnetic Field Measurement Measuring the magnet strength... 15:12:56 Integrated Voltage Measurement... 15:13:18 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = 0.0100000 m VT = -0.021388291 Vs 15:13:33 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = -0.0100000 m VT = 0.021439541 Vs 15:13:48 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = 0.0100000 m VT = -0.021388730 Vs 15:14:03 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = -0.0100000 m VT = 0.021433812 Vs 15:14:18 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = 0.0100000 m VT = -0.021395916 Vs 15:14:33 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = -0.0100000 m VT = 0.021423334 Vs 15:14:48 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = 0.0100000 m VT = -0.021394964 Vs 15:15:03 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = -0.0100000 m VT = 0.021426336 Vs 15:15:04 Integrated Voltage Average and RMS of Measurements X0 = -0.0200000 m, Delta X = 0.0100000 m VT = -0.021411366 +- 0.000001586 Vs 15:15:04 Integrated Field Strength, Average and RMS of Measurements X0 = -0.0200000 m, dx = 0.0100000 m, Nturns = 10.0 VT = -0.021411366 +- 0.000001586 Vs BL = -0.21411366 +- 0.00001586 Tm 15:15:04 Integrated Voltage Measurement... 15:15:22 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = 0.0100000 m VT = -0.010972937 Vs 15:15:37 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = -0.0100000 m VT = 0.011037252 Vs 15:15:52 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = 0.0100000 m VT = -0.010966360 Vs 15:16:07 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = -0.0100000 m VT = 0.011028636 Vs 15:16:23 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = 0.0100000 m VT = -0.010972778 Vs 15:16:38 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = -0.0100000 m VT = 0.011013836 Vs 15:16:53 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = 0.0100000 m VT = -0.010995041 Vs 15:17:08 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = -0.0100000 m VT = 0.011008905 Vs 15:17:08 Integrated Voltage Average and RMS of Measurements X0 = -0.0100000 m, Delta X = 0.0100000 m VT = -0.010999468 +- 0.000004466 Vs 15:17:08 Integrated Field Strength, Average and RMS of Measurements X0 = -0.0100000 m, dx = 0.0100000 m, Nturns = 10.0 VT = -0.010999468 +- 0.000004466 Vs BL = -0.10999468 +- 0.00004466 Tm 15:17:08 Integrated Voltage Measurement... 15:17:26 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = -0.000577628 Vs 15:17:42 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = 0.000601067 Vs 15:17:57 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = -0.000574153 Vs 15:18:12 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = 0.000592235 Vs 15:18:27 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = -0.000582989 Vs 15:18:42 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = 0.000591392 Vs 15:18:57 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = -0.000580390 Vs 15:19:12 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = 0.000600042 Vs 15:19:12 Integrated Voltage Average and RMS of Measurements X0 = 0.0000000 m, Delta X = 0.0100000 m VT = -0.000587487 +- 0.000002713 Vs 15:19:12 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0000000 m, dx = 0.0100000 m, Nturns = 10.0 VT = -0.000587487 +- 0.000002713 Vs BL = -0.00587487 +- 0.00002713 Tm 15:19:12 Integrated Voltage Measurement... 15:19:31 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = 0.0100000 m VT = 0.009825834 Vs 15:19:46 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = -0.0100000 m VT = -0.009809824 Vs 15:20:01 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = 0.0100000 m VT = 0.009829410 Vs 15:20:16 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = -0.0100000 m VT = -0.009798048 Vs 15:20:31 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = 0.0100000 m VT = 0.009838835 Vs 15:20:46 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = -0.0100000 m VT = -0.009805142 Vs 15:21:01 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = 0.0100000 m VT = 0.009827989 Vs 15:21:16 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = -0.0100000 m VT = -0.009799543 Vs 15:21:16 Integrated Voltage Average and RMS of Measurements X0 = 0.0100000 m, Delta X = 0.0100000 m VT = 0.009816828 +- 0.000003414 Vs 15:21:16 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0100000 m, dx = 0.0100000 m, Nturns = 10.0 VT = 0.009816828 +- 0.000003414 Vs BL = 0.09816828 +- 0.00003414 Tm 15:21:16 Integrated Voltage Measurement... 15:21:35 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = 0.0100000 m VT = 0.020237678 Vs 15:21:50 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = -0.0100000 m VT = -0.020203399 Vs 15:22:05 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = 0.0100000 m VT = 0.020228776 Vs 15:22:20 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = -0.0100000 m VT = -0.020209353 Vs 15:22:35 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = 0.0100000 m VT = 0.020234757 Vs 15:22:50 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = -0.0100000 m VT = -0.020203527 Vs 15:23:05 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = 0.0100000 m VT = 0.020230064 Vs 15:23:20 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = -0.0100000 m VT = -0.020205278 Vs 15:23:20 Integrated Voltage Average and RMS of Measurements X0 = 0.0200000 m, Delta X = 0.0100000 m VT = 0.020219104 +- 0.000001014 Vs 15:23:20 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0200000 m, dx = 0.0100000 m, Nturns = 10.0 VT = 0.020219104 +- 0.000001014 Vs BL = 0.20219104 +- 0.00001014 Tm 15:23:25 BL vs X Fit Coefficients: Cn sigCn n (Tm/m^n) ----- --------+-------- 0 -0.00592 0.00001 1 10.40772 0.00068 Compute Fit Residuals: X0 BL sigBL BLfit BL-BLfit (m) (Tm) (Tm) (Tm) (Tm) ---------- --------+-------- -------- -------- -0.020000 -0.21411 0.00002 -0.21408 -0.00003 -0.010000 -0.10999 0.00004 -0.11000 0.00001 0.000000 -0.00587 0.00003 -0.00592 0.00005 0.010000 0.09817 0.00003 0.09815 0.00002 0.020000 0.20219 0.00001 0.20223 -0.00004 15:23:25 Magnet Strength Calculation Harmonic Number = 2 Number Of X0 Positions = 5 X0 BL sigBL (m) (Tm) (Tm) -------- --------+-------- -0.0200 -0.21411 0.00002 -0.0100 -0.10999 0.00004 0.0000 -0.00587 0.00003 0.0100 0.09817 0.00003 0.0200 0.20219 0.00001 SL = 10.4077236 +- 0.0006765 Tm^2/m^n 15:23:25 Quadrupole Strength Calculation Check Slope = 10.407724 Tm/m, Inter = -0.005925 Tm, Sig_slope = 0.001351 Tm/m, Est_sig = 0.000043 Tm GL = 10.407724 +- 0.001351 T 15:23:25 Magnet Strength Measurement Harmonic Number = 2 SL = 10.4077236 +- 0.0006765 Tm^2/m^n 15:23:27 Magnet Current, Imag = 80.5004 A Vtrans = 0.805004 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 15:23:28 Magnet Current, Imag = 80.5006 A Vtrans = 0.805006 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 15:23:29 Magnet Current, Imag = 80.5005 A Vtrans = 0.805005 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 15:23:31 Magnet Current, Imag = 80.5008 A Vtrans = 0.805008 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 15:23:31 Magnet Current Measurement: Imag = 80.5006 +- 0.0001 A 15:23:31 Magnet current ramping to 85.0000 A... 15:23:39 Magnet Current, Imag = 85.4758 A Vtrans = 0.854758 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 15:23:41 Magnet Current, Imag = 85.4760 A Vtrans = 0.854760 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 15:23:42 Magnet Current, Imag = 85.4750 A Vtrans = 0.854750 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 15:23:43 Magnet Current, Imag = 85.4755 A Vtrans = 0.854755 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 15:23:43 Magnet Current Measurement: Imag = 85.4756 +- 0.0004 A 15:23:43 Magnetic Field Measurement Measuring the magnet strength... 15:23:43 Integrated Voltage Measurement... 15:24:06 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = 0.0100000 m VT = -0.022565804 Vs 15:24:21 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = -0.0100000 m VT = 0.022639027 Vs 15:24:36 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = 0.0100000 m VT = -0.022566260 Vs 15:24:51 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = -0.0100000 m VT = 0.022632951 Vs 15:25:07 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = 0.0100000 m VT = -0.022565022 Vs 15:25:22 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = -0.0100000 m VT = 0.022636303 Vs 15:25:37 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = 0.0100000 m VT = -0.022571316 Vs 15:25:52 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = -0.0100000 m VT = 0.022631224 Vs 15:25:52 Integrated Voltage Average and RMS of Measurements X0 = -0.0200000 m, Delta X = 0.0100000 m VT = -0.022600988 +- 0.000001017 Vs 15:25:52 Integrated Field Strength, Average and RMS of Measurements X0 = -0.0200000 m, dx = 0.0100000 m, Nturns = 10.0 VT = -0.022600988 +- 0.000001017 Vs BL = -0.22600988 +- 0.00001017 Tm 15:25:52 Integrated Voltage Measurement... 15:26:10 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = 0.0100000 m VT = -0.011585035 Vs 15:26:25 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = -0.0100000 m VT = 0.011646353 Vs 15:26:40 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = 0.0100000 m VT = -0.011575906 Vs 15:26:55 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = -0.0100000 m VT = 0.011640957 Vs 15:27:10 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = 0.0100000 m VT = -0.011567381 Vs 15:27:26 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = -0.0100000 m VT = 0.011645634 Vs 15:27:41 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = 0.0100000 m VT = -0.011577685 Vs 15:27:56 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = -0.0100000 m VT = 0.011644670 Vs 15:27:56 Integrated Voltage Average and RMS of Measurements X0 = -0.0100000 m, Delta X = 0.0100000 m VT = -0.011610453 +- 0.000003452 Vs 15:27:56 Integrated Field Strength, Average and RMS of Measurements X0 = -0.0100000 m, dx = 0.0100000 m, Nturns = 10.0 VT = -0.011610453 +- 0.000003452 Vs BL = -0.11610453 +- 0.00003452 Tm 15:27:56 Integrated Voltage Measurement... 15:28:14 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = -0.000601070 Vs 15:28:29 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = 0.000638552 Vs 15:28:44 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = -0.000599379 Vs 15:29:00 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = 0.000633493 Vs 15:29:15 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = -0.000611721 Vs 15:29:30 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = 0.000636646 Vs 15:29:45 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = -0.000593476 Vs 15:30:00 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = 0.000649793 Vs 15:30:00 Integrated Voltage Average and RMS of Measurements X0 = 0.0000000 m, Delta X = 0.0100000 m VT = -0.000620516 +- 0.000002821 Vs 15:30:00 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0000000 m, dx = 0.0100000 m, Nturns = 10.0 VT = -0.000620516 +- 0.000002821 Vs BL = -0.00620516 +- 0.00002821 Tm 15:30:00 Integrated Voltage Measurement... 15:30:18 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = 0.0100000 m VT = 0.010391261 Vs 15:30:34 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = -0.0100000 m VT = -0.010345234 Vs 15:30:49 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = 0.0100000 m VT = 0.010379196 Vs 15:31:04 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = -0.0100000 m VT = -0.010348414 Vs 15:31:19 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = 0.0100000 m VT = 0.010377387 Vs 15:31:34 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = -0.0100000 m VT = -0.010343138 Vs 15:31:49 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = 0.0100000 m VT = 0.010380351 Vs 15:32:04 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = -0.0100000 m VT = -0.010343832 Vs 15:32:04 Integrated Voltage Average and RMS of Measurements X0 = 0.0100000 m, Delta X = 0.0100000 m VT = 0.010363601 +- 0.000002960 Vs 15:32:04 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0100000 m, dx = 0.0100000 m, Nturns = 10.0 VT = 0.010363601 +- 0.000002960 Vs BL = 0.10363601 +- 0.00002960 Tm 15:32:04 Integrated Voltage Measurement... 15:32:23 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = 0.0100000 m VT = 0.021365132 Vs 15:32:38 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = -0.0100000 m VT = -0.021317054 Vs 15:32:53 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = 0.0100000 m VT = 0.021368433 Vs 15:33:08 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = -0.0100000 m VT = -0.021320699 Vs 15:33:23 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = 0.0100000 m VT = 0.021362239 Vs 15:33:38 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = -0.0100000 m VT = -0.021321622 Vs 15:33:53 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = 0.0100000 m VT = 0.021355336 Vs 15:34:08 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = -0.0100000 m VT = -0.021331068 Vs 15:34:08 Integrated Voltage Average and RMS of Measurements X0 = 0.0200000 m, Delta X = 0.0100000 m VT = 0.021342698 +- 0.000001314 Vs 15:34:08 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0200000 m, dx = 0.0100000 m, Nturns = 10.0 VT = 0.021342698 +- 0.000001314 Vs BL = 0.21342698 +- 0.00001314 Tm 15:34:14 BL vs X Fit Coefficients: Cn sigCn n (Tm/m^n) ----- --------+-------- 0 -0.00625 0.00001 1 10.98614 0.00056 Compute Fit Residuals: X0 BL sigBL BLfit BL-BLfit (m) (Tm) (Tm) (Tm) (Tm) ---------- --------+-------- -------- -------- -0.020000 -0.22601 0.00001 -0.22597 -0.00004 -0.010000 -0.11610 0.00003 -0.11611 0.00001 0.000000 -0.00621 0.00003 -0.00625 0.00005 0.010000 0.10364 0.00003 0.10361 0.00003 0.020000 0.21343 0.00001 0.21347 -0.00004 15:34:14 Magnet Strength Calculation Harmonic Number = 2 Number Of X0 Positions = 5 X0 BL sigBL (m) (Tm) (Tm) -------- --------+-------- -0.0200 -0.22601 0.00001 -0.0100 -0.11610 0.00003 0.0000 -0.00621 0.00003 0.0100 0.10364 0.00003 0.0200 0.21343 0.00001 SL = 10.9861427 +- 0.0005632 Tm^2/m^n 15:34:14 Quadrupole Strength Calculation Check Slope = 10.986143 Tm/m, Inter = -0.006251 Tm, Sig_slope = 0.001429 Tm/m, Est_sig = 0.000045 Tm GL = 10.986143 +- 0.001429 T 15:34:14 Magnet Strength Measurement Harmonic Number = 2 SL = 10.9861427 +- 0.0005632 Tm^2/m^n 15:34:15 Magnet Current, Imag = 85.4792 A Vtrans = 0.854792 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 15:34:16 Magnet Current, Imag = 85.4794 A Vtrans = 0.854794 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 15:34:18 Magnet Current, Imag = 85.4781 A Vtrans = 0.854781 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 15:34:19 Magnet Current, Imag = 85.4788 A Vtrans = 0.854788 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 15:34:19 Magnet Current Measurement: Imag = 85.4789 +- 0.0005 A 15:34:19 Magnet current ramping to 90.0000 A... 15:34:28 Magnet Current, Imag = 90.5425 A Vtrans = 0.905425 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 15:34:29 Magnet Current, Imag = 90.5423 A Vtrans = 0.905423 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 15:34:30 Magnet Current, Imag = 90.5419 A Vtrans = 0.905419 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 15:34:32 Magnet Current, Imag = 90.5416 A Vtrans = 0.905416 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 15:34:32 Magnet Current Measurement: Imag = 90.5421 +- 0.0003 A 15:34:32 Magnetic Field Measurement Measuring the magnet strength... 15:34:32 Integrated Voltage Measurement... 15:34:54 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = 0.0100000 m VT = -0.023694616 Vs 15:35:09 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = -0.0100000 m VT = 0.023715352 Vs 15:35:24 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = 0.0100000 m VT = -0.023675521 Vs 15:35:39 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = -0.0100000 m VT = 0.023717890 Vs 15:35:54 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = 0.0100000 m VT = -0.023682046 Vs 15:36:10 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = -0.0100000 m VT = 0.023728501 Vs 15:36:25 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = 0.0100000 m VT = -0.023674453 Vs 15:36:40 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = -0.0100000 m VT = 0.023726413 Vs 15:36:40 Integrated Voltage Average and RMS of Measurements X0 = -0.0200000 m, Delta X = 0.0100000 m VT = -0.023701849 +- 0.000003536 Vs 15:36:40 Integrated Field Strength, Average and RMS of Measurements X0 = -0.0200000 m, dx = 0.0100000 m, Nturns = 10.0 VT = -0.023701849 +- 0.000003536 Vs BL = -0.23701849 +- 0.00003536 Tm 15:36:40 Integrated Voltage Measurement... 15:36:58 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = 0.0100000 m VT = -0.012152273 Vs 15:37:13 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = -0.0100000 m VT = 0.012200357 Vs 15:37:29 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = 0.0100000 m VT = -0.012152564 Vs 15:37:44 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = -0.0100000 m VT = 0.012192435 Vs 15:37:59 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = 0.0100000 m VT = -0.012146607 Vs 15:38:14 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = -0.0100000 m VT = 0.012196071 Vs 15:38:29 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = 0.0100000 m VT = -0.012148986 Vs 15:38:44 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = -0.0100000 m VT = 0.012213059 Vs 15:38:44 Integrated Voltage Average and RMS of Measurements X0 = -0.0100000 m, Delta X = 0.0100000 m VT = -0.012175294 +- 0.000003785 Vs 15:38:44 Integrated Field Strength, Average and RMS of Measurements X0 = -0.0100000 m, dx = 0.0100000 m, Nturns = 10.0 VT = -0.012175294 +- 0.000003785 Vs BL = -0.12175294 +- 0.00003785 Tm 15:38:44 Integrated Voltage Measurement... 15:39:03 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = -0.000615307 Vs 15:39:18 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = 0.000673880 Vs 15:39:36 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = -0.000628821 Vs 15:39:51 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = 0.000671669 Vs 15:40:06 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = -0.000638074 Vs 15:40:22 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = 0.000661206 Vs 15:40:37 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = -0.000628660 Vs 15:40:52 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = 0.000663295 Vs 15:40:52 Integrated Voltage Average and RMS of Measurements X0 = 0.0000000 m, Delta X = 0.0100000 m VT = -0.000647614 +- 0.000002389 Vs 15:40:52 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0000000 m, dx = 0.0100000 m, Nturns = 10.0 VT = -0.000647614 +- 0.000002389 Vs BL = -0.00647614 +- 0.00002389 Tm 15:40:52 Integrated Voltage Measurement... 15:41:10 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = 0.0100000 m VT = 0.010885697 Vs 15:41:25 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = -0.0100000 m VT = -0.010848031 Vs 15:41:40 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = 0.0100000 m VT = 0.010888920 Vs 15:41:55 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = -0.0100000 m VT = -0.010846122 Vs 15:42:10 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = 0.0100000 m VT = 0.010891894 Vs 15:42:25 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = -0.0100000 m VT = -0.010847496 Vs 15:42:41 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = 0.0100000 m VT = 0.010880275 Vs 15:42:56 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = -0.0100000 m VT = -0.010867008 Vs 15:42:56 Integrated Voltage Average and RMS of Measurements X0 = 0.0100000 m, Delta X = 0.0100000 m VT = 0.010869430 +- 0.000002647 Vs 15:42:56 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0100000 m, dx = 0.0100000 m, Nturns = 10.0 VT = 0.010869430 +- 0.000002647 Vs BL = 0.10869430 +- 0.00002647 Tm 15:42:56 Integrated Voltage Measurement... 15:43:14 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = 0.0100000 m VT = 0.022369738 Vs 15:43:29 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = -0.0100000 m VT = -0.022392998 Vs 15:43:44 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = 0.0100000 m VT = 0.022373577 Vs 15:43:59 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = -0.0100000 m VT = -0.022396345 Vs 15:44:14 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = 0.0100000 m VT = 0.022365676 Vs 15:44:29 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = -0.0100000 m VT = -0.022400361 Vs 15:44:44 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = 0.0100000 m VT = 0.022368051 Vs 15:45:00 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = -0.0100000 m VT = -0.022401190 Vs 15:45:00 Integrated Voltage Average and RMS of Measurements X0 = 0.0200000 m, Delta X = 0.0100000 m VT = 0.022383492 +- 0.000001429 Vs 15:45:00 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0200000 m, dx = 0.0100000 m, Nturns = 10.0 VT = 0.022383492 +- 0.000001429 Vs BL = 0.22383492 +- 0.00001429 Tm 15:45:05 BL vs X Fit Coefficients: Cn sigCn n (Tm/m^n) ----- --------+-------- 0 -0.00654 0.00001 1 11.52154 0.00089 Compute Fit Residuals: X0 BL sigBL BLfit BL-BLfit (m) (Tm) (Tm) (Tm) (Tm) ---------- --------+-------- -------- -------- -0.020000 -0.23702 0.00004 -0.23697 -0.00004 -0.010000 -0.12175 0.00004 -0.12176 0.00001 0.000000 -0.00648 0.00002 -0.00654 0.00007 0.010000 0.10869 0.00003 0.10867 0.00002 0.020000 0.22383 0.00001 0.22389 -0.00005 15:45:05 Magnet Strength Calculation Harmonic Number = 2 Number Of X0 Positions = 5 X0 BL sigBL (m) (Tm) (Tm) -------- --------+-------- -0.0200 -0.23702 0.00004 -0.0100 -0.12175 0.00004 0.0000 -0.00648 0.00002 0.0100 0.10869 0.00003 0.0200 0.22383 0.00001 SL = 11.5215406 +- 0.0008917 Tm^2/m^n 15:45:05 Quadrupole Strength Calculation Check Slope = 11.521541 Tm/m, Inter = -0.006544 Tm, Sig_slope = 0.001805 Tm/m, Est_sig = 0.000057 Tm GL = 11.521541 +- 0.001805 T 15:45:05 Magnet Strength Measurement Harmonic Number = 2 SL = 11.5215406 +- 0.0008917 Tm^2/m^n 15:45:06 Magnet Current, Imag = 90.5459 A Vtrans = 0.905459 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 15:45:08 Magnet Current, Imag = 90.5453 A Vtrans = 0.905453 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 15:45:09 Magnet Current, Imag = 90.5455 A Vtrans = 0.905455 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 15:45:10 Magnet Current, Imag = 90.5459 A Vtrans = 0.905459 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 15:45:10 Magnet Current Measurement: Imag = 90.5457 +- 0.0003 A 15:45:10 Magnet current ramping to 80.0000 A... 15:45:22 Magnet Current, Imag = 80.5063 A Vtrans = 0.805063 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 15:45:23 Magnet Current, Imag = 80.5074 A Vtrans = 0.805074 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 15:45:24 Magnet Current, Imag = 80.5067 A Vtrans = 0.805067 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 15:45:26 Magnet Current, Imag = 80.5066 A Vtrans = 0.805066 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 15:45:26 Magnet Current Measurement: Imag = 80.5067 +- 0.0004 A 15:45:26 Magnetic Field Measurement Measuring the magnet strength... 15:45:26 Integrated Voltage Measurement... 15:45:48 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = 0.0100000 m VT = -0.021533094 Vs 15:46:03 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = -0.0100000 m VT = 0.021515978 Vs 15:46:18 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = 0.0100000 m VT = -0.021517633 Vs 15:46:33 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = -0.0100000 m VT = 0.021522455 Vs 15:46:48 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = 0.0100000 m VT = -0.021519186 Vs 15:47:03 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = -0.0100000 m VT = 0.021521969 Vs 15:47:18 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = 0.0100000 m VT = -0.021509509 Vs 15:47:34 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = -0.0100000 m VT = 0.021510281 Vs 15:47:34 Integrated Voltage Average and RMS of Measurements X0 = -0.0200000 m, Delta X = 0.0100000 m VT = -0.021518763 +- 0.000005406 Vs 15:47:34 Integrated Field Strength, Average and RMS of Measurements X0 = -0.0200000 m, dx = 0.0100000 m, Nturns = 10.0 VT = -0.021518763 +- 0.000005406 Vs BL = -0.21518763 +- 0.00005406 Tm 15:47:34 Integrated Voltage Measurement... 15:47:52 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = 0.0100000 m VT = -0.011062730 Vs 15:48:07 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = -0.0100000 m VT = 0.011055831 Vs 15:48:22 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = 0.0100000 m VT = -0.011067206 Vs 15:48:37 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = -0.0100000 m VT = 0.011050416 Vs 15:48:52 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = 0.0100000 m VT = -0.011057318 Vs 15:49:07 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = -0.0100000 m VT = 0.011050663 Vs 15:49:22 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = 0.0100000 m VT = -0.011049953 Vs 15:49:38 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = -0.0100000 m VT = 0.011054364 Vs 15:49:38 Integrated Voltage Average and RMS of Measurements X0 = -0.0100000 m, Delta X = 0.0100000 m VT = -0.011056060 +- 0.000003059 Vs 15:49:38 Integrated Field Strength, Average and RMS of Measurements X0 = -0.0100000 m, dx = 0.0100000 m, Nturns = 10.0 VT = -0.011056060 +- 0.000003059 Vs BL = -0.11056060 +- 0.00003059 Tm 15:49:38 Integrated Voltage Measurement... 15:49:56 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = -0.000575293 Vs 15:50:11 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = 0.000595306 Vs 15:50:26 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = -0.000587825 Vs 15:50:41 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = 0.000601178 Vs 15:50:57 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = -0.000581487 Vs 15:51:12 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = 0.000597612 Vs 15:51:27 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = -0.000591843 Vs 15:51:42 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = 0.000580931 Vs 15:51:42 Integrated Voltage Average and RMS of Measurements X0 = 0.0000000 m, Delta X = 0.0100000 m VT = -0.000588934 +- 0.000003573 Vs 15:51:42 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0000000 m, dx = 0.0100000 m, Nturns = 10.0 VT = -0.000588934 +- 0.000003573 Vs BL = -0.00588934 +- 0.00003573 Tm 15:51:42 Integrated Voltage Measurement... 15:52:00 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = 0.0100000 m VT = 0.009854787 Vs 15:52:15 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = -0.0100000 m VT = -0.009880543 Vs 15:52:30 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = 0.0100000 m VT = 0.009855215 Vs 15:52:45 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = -0.0100000 m VT = -0.009875193 Vs 15:53:00 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = 0.0100000 m VT = 0.009855005 Vs 15:53:15 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = -0.0100000 m VT = -0.009861323 Vs 15:53:31 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = 0.0100000 m VT = 0.009864831 Vs 15:53:46 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = -0.0100000 m VT = -0.009861921 Vs 15:53:46 Integrated Voltage Average and RMS of Measurements X0 = 0.0100000 m, Delta X = 0.0100000 m VT = 0.009863602 +- 0.000003489 Vs 15:53:46 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0100000 m, dx = 0.0100000 m, Nturns = 10.0 VT = 0.009863602 +- 0.000003489 Vs BL = 0.09863602 +- 0.00003489 Tm 15:53:46 Integrated Voltage Measurement... 15:54:04 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = 0.0100000 m VT = 0.020324899 Vs 15:54:19 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = -0.0100000 m VT = -0.020314143 Vs 15:54:34 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = 0.0100000 m VT = 0.020340308 Vs 15:54:49 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = -0.0100000 m VT = -0.020306630 Vs 15:55:04 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = 0.0100000 m VT = 0.020333765 Vs 15:55:19 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = -0.0100000 m VT = -0.020300594 Vs 15:55:34 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = 0.0100000 m VT = 0.020334771 Vs 15:55:49 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = -0.0100000 m VT = -0.020298387 Vs 15:55:50 Integrated Voltage Average and RMS of Measurements X0 = 0.0200000 m, Delta X = 0.0100000 m VT = 0.020319187 +- 0.000002705 Vs 15:55:50 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0200000 m, dx = 0.0100000 m, Nturns = 10.0 VT = 0.020319187 +- 0.000002705 Vs BL = 0.20319187 +- 0.00002705 Tm 15:55:55 BL vs X Fit Coefficients: Cn sigCn n (Tm/m^n) ----- --------+-------- 0 -0.00596 0.00002 1 10.45956 0.00130 Compute Fit Residuals: X0 BL sigBL BLfit BL-BLfit (m) (Tm) (Tm) (Tm) (Tm) ---------- --------+-------- -------- -------- -0.020000 -0.21519 0.00005 -0.21515 -0.00003 -0.010000 -0.11056 0.00003 -0.11056 -0.00000 0.000000 -0.00589 0.00004 -0.00596 0.00007 0.010000 0.09864 0.00003 0.09863 0.00000 0.020000 0.20319 0.00003 0.20323 -0.00004 15:55:55 Magnet Strength Calculation Harmonic Number = 2 Number Of X0 Positions = 5 X0 BL sigBL (m) (Tm) (Tm) -------- --------+-------- -0.0200 -0.21519 0.00005 -0.0100 -0.11056 0.00003 0.0000 -0.00589 0.00004 0.0100 0.09864 0.00003 0.0200 0.20319 0.00003 SL = 10.4595563 +- 0.0012950 Tm^2/m^n 15:55:55 Quadrupole Strength Calculation Check Slope = 10.459556 Tm/m, Inter = -0.005962 Tm, Sig_slope = 0.001620 Tm/m, Est_sig = 0.000051 Tm GL = 10.459556 +- 0.001620 T 15:55:55 Magnet Strength Measurement Harmonic Number = 2 SL = 10.4595563 +- 0.0012950 Tm^2/m^n 15:55:56 Magnet Current, Imag = 80.5018 A Vtrans = 0.805018 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 15:55:58 Magnet Current, Imag = 80.5014 A Vtrans = 0.805014 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 15:55:59 Magnet Current, Imag = 80.5025 A Vtrans = 0.805025 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 15:56:00 Magnet Current, Imag = 80.5024 A Vtrans = 0.805024 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 15:56:00 Magnet Current Measurement: Imag = 80.5020 +- 0.0004 A 15:56:00 Magnet current ramping to 70.0000 A... 15:56:12 Magnet Current, Imag = 70.3705 A Vtrans = 0.703705 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 15:56:13 Magnet Current, Imag = 70.3702 A Vtrans = 0.703702 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 15:56:14 Magnet Current, Imag = 70.3706 A Vtrans = 0.703706 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 15:56:16 Magnet Current, Imag = 70.3707 A Vtrans = 0.703707 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 15:56:16 Magnet Current Measurement: Imag = 70.3705 +- 0.0002 A 15:56:16 Magnetic Field Measurement Measuring the magnet strength... 15:56:16 Integrated Voltage Measurement... 15:56:38 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = 0.0100000 m VT = -0.018912272 Vs 15:56:53 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = -0.0100000 m VT = 0.018914411 Vs 15:57:08 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = 0.0100000 m VT = -0.018934858 Vs 15:57:23 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = -0.0100000 m VT = 0.018904550 Vs 15:57:38 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = 0.0100000 m VT = -0.018931917 Vs 15:57:53 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = -0.0100000 m VT = 0.018901679 Vs 15:58:08 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = 0.0100000 m VT = -0.018938945 Vs 15:58:23 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = -0.0100000 m VT = 0.018908138 Vs 15:58:23 Integrated Voltage Average and RMS of Measurements X0 = -0.0200000 m, Delta X = 0.0100000 m VT = -0.018918346 +- 0.000003751 Vs 15:58:23 Integrated Field Strength, Average and RMS of Measurements X0 = -0.0200000 m, dx = 0.0100000 m, Nturns = 10.0 VT = -0.018918346 +- 0.000003751 Vs BL = -0.18918346 +- 0.00003751 Tm 15:58:23 Integrated Voltage Measurement... 15:58:42 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = 0.0100000 m VT = -0.009729979 Vs 15:58:57 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = -0.0100000 m VT = 0.009714805 Vs 15:59:12 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = 0.0100000 m VT = -0.009721289 Vs 15:59:27 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = -0.0100000 m VT = 0.009710945 Vs 15:59:43 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = 0.0100000 m VT = -0.009733284 Vs 15:59:58 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = -0.0100000 m VT = 0.009691996 Vs 16:00:13 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = 0.0100000 m VT = -0.009728653 Vs 16:00:28 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = -0.0100000 m VT = 0.009712887 Vs 16:00:28 Integrated Voltage Average and RMS of Measurements X0 = -0.0100000 m, Delta X = 0.0100000 m VT = -0.009717980 +- 0.000003848 Vs 16:00:28 Integrated Field Strength, Average and RMS of Measurements X0 = -0.0100000 m, dx = 0.0100000 m, Nturns = 10.0 VT = -0.009717980 +- 0.000003848 Vs BL = -0.09717980 +- 0.00003848 Tm 16:00:28 Integrated Voltage Measurement... 16:00:46 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = -0.000520860 Vs 16:01:01 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = 0.000528166 Vs 16:01:16 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = -0.000513511 Vs 16:01:31 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = 0.000542852 Vs 16:01:46 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = -0.000501901 Vs 16:02:01 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = 0.000533681 Vs 16:02:17 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = -0.000513014 Vs 16:02:32 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = 0.000528000 Vs 16:02:32 Integrated Voltage Average and RMS of Measurements X0 = 0.0000000 m, Delta X = 0.0100000 m VT = -0.000522748 +- 0.000003944 Vs 16:02:32 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0000000 m, dx = 0.0100000 m, Nturns = 10.0 VT = -0.000522748 +- 0.000003944 Vs BL = -0.00522748 +- 0.00003944 Tm 16:02:32 Integrated Voltage Measurement... 16:02:50 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = 0.0100000 m VT = 0.008674542 Vs 16:03:05 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = -0.0100000 m VT = -0.008669948 Vs 16:03:20 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = 0.0100000 m VT = 0.008673023 Vs 16:03:35 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = -0.0100000 m VT = -0.008670489 Vs 16:03:50 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = 0.0100000 m VT = 0.008673761 Vs 16:04:05 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = -0.0100000 m VT = -0.008674501 Vs 16:04:22 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = 0.0100000 m VT = 0.008669518 Vs 16:04:37 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = -0.0100000 m VT = -0.008671216 Vs 16:04:37 Integrated Voltage Average and RMS of Measurements X0 = 0.0100000 m, Delta X = 0.0100000 m VT = 0.008672125 +- 0.000001348 Vs 16:04:37 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0100000 m, dx = 0.0100000 m, Nturns = 10.0 VT = 0.008672125 +- 0.000001348 Vs BL = 0.08672125 +- 0.00001348 Tm 16:04:37 Integrated Voltage Measurement... 16:04:56 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = 0.0100000 m VT = 0.017866823 Vs 16:05:11 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = -0.0100000 m VT = -0.017852403 Vs 16:05:26 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = 0.0100000 m VT = 0.017865444 Vs 16:05:41 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = -0.0100000 m VT = -0.017857912 Vs 16:05:56 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = 0.0100000 m VT = 0.017872564 Vs 16:06:11 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = -0.0100000 m VT = -0.017853631 Vs 16:06:26 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = 0.0100000 m VT = 0.017875495 Vs 16:06:41 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = -0.0100000 m VT = -0.017855674 Vs 16:06:41 Integrated Voltage Average and RMS of Measurements X0 = 0.0200000 m, Delta X = 0.0100000 m VT = 0.017862493 +- 0.000002173 Vs 16:06:41 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0200000 m, dx = 0.0100000 m, Nturns = 10.0 VT = 0.017862493 +- 0.000002173 Vs BL = 0.17862493 +- 0.00002173 Tm 16:06:47 BL vs X Fit Coefficients: Cn sigCn n (Tm/m^n) ----- --------+-------- 0 -0.00525 0.00001 1 9.19518 0.00096 Compute Fit Residuals: X0 BL sigBL BLfit BL-BLfit (m) (Tm) (Tm) (Tm) (Tm) ---------- --------+-------- -------- -------- -0.020000 -0.18918 0.00004 -0.18915 -0.00003 -0.010000 -0.09718 0.00004 -0.09720 0.00002 0.000000 -0.00523 0.00004 -0.00525 0.00002 0.010000 0.08672 0.00001 0.08670 0.00002 0.020000 0.17862 0.00002 0.17865 -0.00003 16:06:47 Magnet Strength Calculation Harmonic Number = 2 Number Of X0 Positions = 5 X0 BL sigBL (m) (Tm) (Tm) -------- --------+-------- -0.0200 -0.18918 0.00004 -0.0100 -0.09718 0.00004 0.0000 -0.00523 0.00004 0.0100 0.08672 0.00001 0.0200 0.17862 0.00002 SL = 9.1951784 +- 0.0009581 Tm^2/m^n 16:06:47 Quadrupole Strength Calculation Check Slope = 9.195178 Tm/m, Inter = -0.005249 Tm, Sig_slope = 0.001013 Tm/m, Est_sig = 0.000032 Tm GL = 9.195178 +- 0.001013 T 16:06:47 Magnet Strength Measurement Harmonic Number = 2 SL = 9.1951784 +- 0.0009581 Tm^2/m^n 16:06:48 Magnet Current, Imag = 70.3669 A Vtrans = 0.703669 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 16:06:49 Magnet Current, Imag = 70.3667 A Vtrans = 0.703667 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 16:06:51 Magnet Current, Imag = 70.3664 A Vtrans = 0.703664 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 16:06:52 Magnet Current, Imag = 70.3659 A Vtrans = 0.703659 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 16:06:52 Magnet Current Measurement: Imag = 70.3665 +- 0.0004 A 16:06:52 Magnet current ramping to 60.0000 A... 16:07:03 Magnet Current, Imag = 60.2920 A Vtrans = 0.602920 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 16:07:05 Magnet Current, Imag = 60.2931 A Vtrans = 0.602931 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 16:07:06 Magnet Current, Imag = 60.2929 A Vtrans = 0.602929 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 16:07:07 Magnet Current, Imag = 60.2928 A Vtrans = 0.602928 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 16:07:07 Magnet Current Measurement: Imag = 60.2927 +- 0.0004 A 16:07:07 Magnetic Field Measurement Measuring the magnet strength... 16:07:07 Integrated Voltage Measurement... 16:07:30 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = 0.0100000 m VT = -0.016244218 Vs 16:07:45 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = -0.0100000 m VT = 0.016240132 Vs 16:08:00 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = 0.0100000 m VT = -0.016239950 Vs 16:08:15 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = -0.0100000 m VT = 0.016248419 Vs 16:08:30 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = 0.0100000 m VT = -0.016237145 Vs 16:08:45 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = -0.0100000 m VT = 0.016237367 Vs 16:09:00 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = 0.0100000 m VT = -0.016237649 Vs 16:09:15 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = -0.0100000 m VT = 0.016247205 Vs 16:09:15 Integrated Voltage Average and RMS of Measurements X0 = -0.0200000 m, Delta X = 0.0100000 m VT = -0.016241511 +- 0.000002575 Vs 16:09:15 Integrated Field Strength, Average and RMS of Measurements X0 = -0.0200000 m, dx = 0.0100000 m, Nturns = 10.0 VT = -0.016241511 +- 0.000002575 Vs BL = -0.16241511 +- 0.00002575 Tm 16:09:15 Integrated Voltage Measurement... 16:09:34 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = 0.0100000 m VT = -0.008336345 Vs 16:09:49 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = -0.0100000 m VT = 0.008354483 Vs 16:10:04 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = 0.0100000 m VT = -0.008326011 Vs 16:10:19 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = -0.0100000 m VT = 0.008358054 Vs 16:10:35 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = 0.0100000 m VT = -0.008322086 Vs 16:10:50 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = -0.0100000 m VT = 0.008368750 Vs 16:11:05 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = 0.0100000 m VT = -0.008326907 Vs 16:11:20 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = -0.0100000 m VT = 0.008357410 Vs 16:11:20 Integrated Voltage Average and RMS of Measurements X0 = -0.0100000 m, Delta X = 0.0100000 m VT = -0.008343756 +- 0.000001661 Vs 16:11:20 Integrated Field Strength, Average and RMS of Measurements X0 = -0.0100000 m, dx = 0.0100000 m, Nturns = 10.0 VT = -0.008343756 +- 0.000001661 Vs BL = -0.08343756 +- 0.00001661 Tm 16:11:20 Integrated Voltage Measurement... 16:11:38 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = -0.000429150 Vs 16:11:54 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = 0.000461177 Vs 16:12:09 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = -0.000427521 Vs 16:12:24 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = 0.000477145 Vs 16:12:39 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = -0.000426124 Vs 16:12:54 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = 0.000468359 Vs 16:13:09 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = -0.000430295 Vs 16:13:25 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = 0.000456302 Vs 16:13:25 Integrated Voltage Average and RMS of Measurements X0 = 0.0000000 m, Delta X = 0.0100000 m VT = -0.000447009 +- 0.000003375 Vs 16:13:25 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0000000 m, dx = 0.0100000 m, Nturns = 10.0 VT = -0.000447009 +- 0.000003375 Vs BL = -0.00447009 +- 0.00003375 Tm 16:13:25 Integrated Voltage Measurement... 16:13:43 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = 0.0100000 m VT = 0.007463307 Vs 16:13:58 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = -0.0100000 m VT = -0.007409446 Vs 16:14:13 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = 0.0100000 m VT = 0.007474323 Vs 16:14:28 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = -0.0100000 m VT = -0.007416868 Vs 16:14:43 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = 0.0100000 m VT = 0.007469954 Vs 16:14:58 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = -0.0100000 m VT = -0.007423052 Vs 16:15:13 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = 0.0100000 m VT = 0.007456015 Vs 16:15:29 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = -0.0100000 m VT = -0.007428175 Vs 16:15:29 Integrated Voltage Average and RMS of Measurements X0 = 0.0100000 m, Delta X = 0.0100000 m VT = 0.007442642 +- 0.000003974 Vs 16:15:29 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0100000 m, dx = 0.0100000 m, Nturns = 10.0 VT = 0.007442642 +- 0.000003974 Vs BL = 0.07442642 +- 0.00003974 Tm 16:15:29 Integrated Voltage Measurement... 16:15:47 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = 0.0100000 m VT = 0.015335635 Vs 16:16:02 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = -0.0100000 m VT = -0.015328726 Vs 16:16:17 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = 0.0100000 m VT = 0.015326606 Vs 16:16:32 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = -0.0100000 m VT = -0.015331045 Vs 16:16:47 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = 0.0100000 m VT = 0.015317723 Vs 16:17:02 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = -0.0100000 m VT = -0.015345020 Vs 16:17:17 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = 0.0100000 m VT = 0.015311666 Vs 16:17:32 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = -0.0100000 m VT = -0.015338192 Vs 16:17:32 Integrated Voltage Average and RMS of Measurements X0 = 0.0200000 m, Delta X = 0.0100000 m VT = 0.015329327 +- 0.000002825 Vs 16:17:32 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0200000 m, dx = 0.0100000 m, Nturns = 10.0 VT = 0.015329327 +- 0.000002825 Vs BL = 0.15329327 +- 0.00002825 Tm 16:17:38 BL vs X Fit Coefficients: Cn sigCn n (Tm/m^n) ----- --------+-------- 0 -0.00452 0.00001 1 7.89281 0.00088 Compute Fit Residuals: X0 BL sigBL BLfit BL-BLfit (m) (Tm) (Tm) (Tm) (Tm) ---------- --------+-------- -------- -------- -0.020000 -0.16242 0.00003 -0.16238 -0.00004 -0.010000 -0.08344 0.00002 -0.08345 0.00001 0.000000 -0.00447 0.00003 -0.00452 0.00005 0.010000 0.07443 0.00004 0.07441 0.00002 0.020000 0.15329 0.00003 0.15334 -0.00004 16:17:38 Magnet Strength Calculation Harmonic Number = 2 Number Of X0 Positions = 5 X0 BL sigBL (m) (Tm) (Tm) -------- --------+-------- -0.0200 -0.16242 0.00003 -0.0100 -0.08344 0.00002 0.0000 -0.00447 0.00003 0.0100 0.07443 0.00004 0.0200 0.15329 0.00003 SL = 7.8928072 +- 0.0008775 Tm^2/m^n 16:17:38 Quadrupole Strength Calculation Check Slope = 7.892807 Tm/m, Inter = -0.004521 Tm, Sig_slope = 0.001448 Tm/m, Est_sig = 0.000046 Tm GL = 7.892807 +- 0.001448 T 16:17:38 Magnet Strength Measurement Harmonic Number = 2 SL = 7.8928072 +- 0.0008775 Tm^2/m^n 16:17:39 Magnet Current, Imag = 60.2877 A Vtrans = 0.602877 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 16:17:40 Magnet Current, Imag = 60.2884 A Vtrans = 0.602884 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 16:17:42 Magnet Current, Imag = 60.2881 A Vtrans = 0.602881 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 16:17:43 Magnet Current, Imag = 60.2881 A Vtrans = 0.602881 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 16:17:43 Magnet Current Measurement: Imag = 60.2881 +- 0.0002 A 16:17:43 Magnet current ramping to 50.0000 A... 16:17:55 Magnet Current, Imag = 50.2722 A Vtrans = 0.502722 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 16:17:56 Magnet Current, Imag = 50.2723 A Vtrans = 0.502723 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 16:17:57 Magnet Current, Imag = 50.2727 A Vtrans = 0.502727 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 16:17:59 Magnet Current, Imag = 50.2724 A Vtrans = 0.502724 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 16:17:59 Magnet Current Measurement: Imag = 50.2724 +- 0.0002 A 16:17:59 Magnetic Field Measurement Measuring the magnet strength... 16:17:59 Integrated Voltage Measurement... 16:18:21 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = 0.0100000 m VT = -0.013553278 Vs 16:18:36 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = -0.0100000 m VT = 0.013553469 Vs 16:18:51 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = 0.0100000 m VT = -0.013573244 Vs 16:19:06 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = -0.0100000 m VT = 0.013547742 Vs 16:19:21 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = 0.0100000 m VT = -0.013575585 Vs 16:19:36 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = -0.0100000 m VT = 0.013525912 Vs 16:19:51 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = 0.0100000 m VT = -0.013602936 Vs 16:20:06 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = -0.0100000 m VT = 0.013513165 Vs 16:20:06 Integrated Voltage Average and RMS of Measurements X0 = -0.0200000 m, Delta X = 0.0100000 m VT = -0.013555666 +- 0.000003822 Vs 16:20:06 Integrated Field Strength, Average and RMS of Measurements X0 = -0.0200000 m, dx = 0.0100000 m, Nturns = 10.0 VT = -0.013555666 +- 0.000003822 Vs BL = -0.13555666 +- 0.00003822 Tm 16:20:06 Integrated Voltage Measurement... 16:20:25 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = 0.0100000 m VT = -0.007005576 Vs 16:20:40 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = -0.0100000 m VT = 0.006921565 Vs 16:20:55 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = 0.0100000 m VT = -0.007031133 Vs 16:21:10 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = -0.0100000 m VT = 0.006893443 Vs 16:21:25 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = 0.0100000 m VT = -0.007039307 Vs 16:21:40 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = -0.0100000 m VT = 0.006895114 Vs 16:21:55 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = 0.0100000 m VT = -0.007026779 Vs 16:22:10 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = -0.0100000 m VT = 0.006898898 Vs 16:22:10 Integrated Voltage Average and RMS of Measurements X0 = -0.0100000 m, Delta X = 0.0100000 m VT = -0.006963977 +- 0.000001922 Vs 16:22:10 Integrated Field Strength, Average and RMS of Measurements X0 = -0.0100000 m, dx = 0.0100000 m, Nturns = 10.0 VT = -0.006963977 +- 0.000001922 Vs BL = -0.06963977 +- 0.00001922 Tm 16:22:10 Integrated Voltage Measurement... 16:22:29 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = -0.000443762 Vs 16:22:44 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = 0.000301540 Vs 16:22:59 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = -0.000446843 Vs 16:23:14 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = 0.000310919 Vs 16:23:29 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = -0.000438617 Vs 16:23:44 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = 0.000308004 Vs 16:23:59 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = -0.000466039 Vs 16:24:14 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = 0.000275105 Vs 16:24:14 Integrated Voltage Average and RMS of Measurements X0 = 0.0000000 m, Delta X = 0.0100000 m VT = -0.000373854 +- 0.000003074 Vs 16:24:14 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0000000 m, dx = 0.0100000 m, Nturns = 10.0 VT = -0.000373854 +- 0.000003074 Vs BL = -0.00373854 +- 0.00003074 Tm 16:24:14 Integrated Voltage Measurement... 16:24:33 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = 0.0100000 m VT = 0.006126908 Vs 16:24:48 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = -0.0100000 m VT = -0.006288704 Vs 16:25:03 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = 0.0100000 m VT = 0.006141303 Vs 16:25:18 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = -0.0100000 m VT = -0.006268451 Vs 16:25:33 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = 0.0100000 m VT = 0.006161406 Vs 16:25:48 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = -0.0100000 m VT = -0.006254476 Vs 16:26:03 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = 0.0100000 m VT = 0.006178443 Vs 16:26:18 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = -0.0100000 m VT = -0.006220637 Vs 16:26:18 Integrated Voltage Average and RMS of Measurements X0 = 0.0100000 m, Delta X = 0.0100000 m VT = 0.006205041 +- 0.000003404 Vs 16:26:18 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0100000 m, dx = 0.0100000 m, Nturns = 10.0 VT = 0.006205041 +- 0.000003404 Vs BL = 0.06205041 +- 0.00003404 Tm 16:26:18 Integrated Voltage Measurement... 16:26:37 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = 0.0100000 m VT = 0.012784888 Vs 16:26:52 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = -0.0100000 m VT = -0.012818283 Vs 16:27:08 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = 0.0100000 m VT = 0.012769467 Vs 16:27:23 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = -0.0100000 m VT = -0.012818177 Vs 16:27:38 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = 0.0100000 m VT = 0.012774811 Vs 16:27:53 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = -0.0100000 m VT = -0.012803993 Vs 16:28:08 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = 0.0100000 m VT = 0.012785011 Vs 16:28:23 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = -0.0100000 m VT = -0.012805087 Vs 16:28:23 Integrated Voltage Average and RMS of Measurements X0 = 0.0200000 m, Delta X = 0.0100000 m VT = 0.012794965 +- 0.000004362 Vs 16:28:23 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0200000 m, dx = 0.0100000 m, Nturns = 10.0 VT = 0.012794965 +- 0.000004362 Vs BL = 0.12794965 +- 0.00004362 Tm 16:28:28 BL vs X Fit Coefficients: Cn sigCn n (Tm/m^n) ----- --------+-------- 0 -0.00379 0.00002 1 6.58703 0.00122 Compute Fit Residuals: X0 BL sigBL BLfit BL-BLfit (m) (Tm) (Tm) (Tm) (Tm) ---------- --------+-------- -------- -------- -0.020000 -0.13556 0.00004 -0.13553 -0.00003 -0.010000 -0.06964 0.00002 -0.06966 0.00002 0.000000 -0.00374 0.00003 -0.00379 0.00005 0.010000 0.06205 0.00003 0.06208 -0.00003 0.020000 0.12795 0.00004 0.12795 -0.00000 16:28:28 Magnet Strength Calculation Harmonic Number = 2 Number Of X0 Positions = 5 X0 BL sigBL (m) (Tm) (Tm) -------- --------+-------- -0.0200 -0.13556 0.00004 -0.0100 -0.06964 0.00002 0.0000 -0.00374 0.00003 0.0100 0.06205 0.00003 0.0200 0.12795 0.00004 SL = 6.5870280 +- 0.0012239 Tm^2/m^n 16:28:28 Quadrupole Strength Calculation Check Slope = 6.587028 Tm/m, Inter = -0.003787 Tm, Sig_slope = 0.001238 Tm/m, Est_sig = 0.000039 Tm GL = 6.587028 +- 0.001238 T 16:28:28 Magnet Strength Measurement Harmonic Number = 2 SL = 6.5870280 +- 0.0012239 Tm^2/m^n 16:28:30 Magnet Current, Imag = 50.2709 A Vtrans = 0.502709 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 16:28:31 Magnet Current, Imag = 50.2713 A Vtrans = 0.502713 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 16:28:32 Magnet Current, Imag = 50.2702 A Vtrans = 0.502702 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 16:28:33 Magnet Current, Imag = 50.2704 A Vtrans = 0.502704 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 16:28:33 Magnet Current Measurement: Imag = 50.2707 +- 0.0004 A 16:28:34 Magnet current ramping to 40.0000 A... 16:28:45 Magnet Current, Imag = 40.1684 A Vtrans = 0.401684 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 16:28:46 Magnet Current, Imag = 40.1684 A Vtrans = 0.401684 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 16:28:48 Magnet Current, Imag = 40.1680 A Vtrans = 0.401680 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 16:28:49 Magnet Current, Imag = 40.1685 A Vtrans = 0.401685 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 16:28:49 Magnet Current Measurement: Imag = 40.1683 +- 0.0002 A 16:28:49 Magnetic Field Measurement Measuring the magnet strength... 16:28:49 Integrated Voltage Measurement... 16:29:12 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = 0.0100000 m VT = -0.010861496 Vs 16:29:27 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = -0.0100000 m VT = 0.010826665 Vs 16:29:42 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = 0.0100000 m VT = -0.010853347 Vs 16:29:57 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = -0.0100000 m VT = 0.010833716 Vs 16:30:13 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = 0.0100000 m VT = -0.010864135 Vs 16:30:28 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = -0.0100000 m VT = 0.010833382 Vs 16:30:43 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = 0.0100000 m VT = -0.010865153 Vs 16:30:58 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = -0.0100000 m VT = 0.010833548 Vs 16:30:58 Integrated Voltage Average and RMS of Measurements X0 = -0.0200000 m, Delta X = 0.0100000 m VT = -0.010846430 +- 0.000002640 Vs 16:30:58 Integrated Field Strength, Average and RMS of Measurements X0 = -0.0200000 m, dx = 0.0100000 m, Nturns = 10.0 VT = -0.010846430 +- 0.000002640 Vs BL = -0.10846430 +- 0.00002640 Tm 16:30:58 Integrated Voltage Measurement... 16:31:16 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = 0.0100000 m VT = -0.005583325 Vs 16:31:31 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = -0.0100000 m VT = 0.005562394 Vs 16:31:46 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = 0.0100000 m VT = -0.005587492 Vs 16:32:01 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = -0.0100000 m VT = 0.005568597 Vs 16:32:16 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = 0.0100000 m VT = -0.005581080 Vs 16:32:31 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = -0.0100000 m VT = 0.005563101 Vs 16:32:47 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = 0.0100000 m VT = -0.005580874 Vs 16:33:02 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = -0.0100000 m VT = 0.005554007 Vs 16:33:02 Integrated Voltage Average and RMS of Measurements X0 = -0.0100000 m, Delta X = 0.0100000 m VT = -0.005572609 +- 0.000003761 Vs 16:33:02 Integrated Field Strength, Average and RMS of Measurements X0 = -0.0100000 m, dx = 0.0100000 m, Nturns = 10.0 VT = -0.005572609 +- 0.000003761 Vs BL = -0.05572609 +- 0.00003761 Tm 16:33:02 Integrated Voltage Measurement... 16:33:20 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = -0.000319084 Vs 16:33:35 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = 0.000280314 Vs 16:33:50 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = -0.000320735 Vs 16:34:06 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = 0.000274356 Vs 16:34:21 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = -0.000329339 Vs 16:34:36 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = 0.000274611 Vs 16:34:51 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = -0.000329333 Vs 16:35:06 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = 0.000281323 Vs 16:35:06 Integrated Voltage Average and RMS of Measurements X0 = 0.0000000 m, Delta X = 0.0100000 m VT = -0.000301137 +- 0.000002882 Vs 16:35:06 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0000000 m, dx = 0.0100000 m, Nturns = 10.0 VT = -0.000301137 +- 0.000002882 Vs BL = -0.00301137 +- 0.00002882 Tm 16:35:06 Integrated Voltage Measurement... 16:35:24 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = 0.0100000 m VT = 0.004949024 Vs 16:35:39 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = -0.0100000 m VT = -0.004978277 Vs 16:35:54 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = 0.0100000 m VT = 0.004961023 Vs 16:36:09 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = -0.0100000 m VT = -0.004972354 Vs 16:36:25 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = 0.0100000 m VT = 0.004957211 Vs 16:36:40 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = -0.0100000 m VT = -0.004977157 Vs 16:36:55 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = 0.0100000 m VT = 0.004955618 Vs 16:37:10 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = -0.0100000 m VT = -0.004975120 Vs 16:37:10 Integrated Voltage Average and RMS of Measurements X0 = 0.0100000 m, Delta X = 0.0100000 m VT = 0.004965723 +- 0.000001368 Vs 16:37:10 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0100000 m, dx = 0.0100000 m, Nturns = 10.0 VT = 0.004965723 +- 0.000001368 Vs BL = 0.04965723 +- 0.00001368 Tm 16:37:10 Integrated Voltage Measurement... 16:37:29 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = 0.0100000 m VT = 0.010218713 Vs 16:37:44 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = -0.0100000 m VT = -0.010255369 Vs 16:37:59 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = 0.0100000 m VT = 0.010206341 Vs 16:38:14 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = -0.0100000 m VT = -0.010257141 Vs 16:38:29 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = 0.0100000 m VT = 0.010205916 Vs 16:38:44 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = -0.0100000 m VT = -0.010254690 Vs 16:38:59 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = 0.0100000 m VT = 0.010208122 Vs 16:39:14 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = -0.0100000 m VT = -0.010262063 Vs 16:39:14 Integrated Voltage Average and RMS of Measurements X0 = 0.0200000 m, Delta X = 0.0100000 m VT = 0.010233544 +- 0.000002664 Vs 16:39:14 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0200000 m, dx = 0.0100000 m, Nturns = 10.0 VT = 0.010233544 +- 0.000002664 Vs BL = 0.10233544 +- 0.00002664 Tm 16:39:19 BL vs X Fit Coefficients: Cn sigCn n (Tm/m^n) ----- --------+-------- 0 -0.00304 0.00001 1 5.26983 0.00085 Compute Fit Residuals: X0 BL sigBL BLfit BL-BLfit (m) (Tm) (Tm) (Tm) (Tm) ---------- --------+-------- -------- -------- -0.020000 -0.10846 0.00003 -0.10844 -0.00003 -0.010000 -0.05573 0.00004 -0.05574 0.00001 0.000000 -0.00301 0.00003 -0.00304 0.00003 0.010000 0.04966 0.00001 0.04966 0.00000 0.020000 0.10234 0.00003 0.10235 -0.00002 16:39:19 Magnet Strength Calculation Harmonic Number = 2 Number Of X0 Positions = 5 X0 BL sigBL (m) (Tm) (Tm) -------- --------+-------- -0.0200 -0.10846 0.00003 -0.0100 -0.05573 0.00004 0.0000 -0.00301 0.00003 0.0100 0.04966 0.00001 0.0200 0.10234 0.00003 SL = 5.2698281 +- 0.0008501 Tm^2/m^n 16:39:19 Quadrupole Strength Calculation Check Slope = 5.269828 Tm/m, Inter = -0.003042 Tm, Sig_slope = 0.000850 Tm/m, Est_sig = 0.000027 Tm GL = 5.269828 +- 0.000850 T 16:39:19 Magnet Strength Measurement Harmonic Number = 2 SL = 5.2698281 +- 0.0008501 Tm^2/m^n 16:39:21 Magnet Current, Imag = 40.1698 A Vtrans = 0.401698 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 16:39:22 Magnet Current, Imag = 40.1670 A Vtrans = 0.401670 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 16:39:23 Magnet Current, Imag = 40.1673 A Vtrans = 0.401673 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 16:39:25 Magnet Current, Imag = 40.1673 A Vtrans = 0.401673 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 16:39:25 Magnet Current Measurement: Imag = 40.1679 +- 0.0011 A 16:39:25 Magnet current ramping to 30.0000 A... 16:39:36 Magnet Current, Imag = 30.1036 A Vtrans = 0.301036 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 16:39:38 Magnet Current, Imag = 30.1032 A Vtrans = 0.301032 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 16:39:39 Magnet Current, Imag = 30.1030 A Vtrans = 0.301030 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 16:39:40 Magnet Current, Imag = 30.1029 A Vtrans = 0.301029 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 16:39:40 Magnet Current Measurement: Imag = 30.1032 +- 0.0003 A 16:39:40 Magnetic Field Measurement Measuring the magnet strength... 16:39:40 Integrated Voltage Measurement... 16:40:02 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = 0.0100000 m VT = -0.008161057 Vs 16:40:18 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = -0.0100000 m VT = 0.008129654 Vs 16:40:33 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = 0.0100000 m VT = -0.008169652 Vs 16:40:48 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = -0.0100000 m VT = 0.008125822 Vs 16:41:03 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = 0.0100000 m VT = -0.008165341 Vs 16:41:18 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = -0.0100000 m VT = 0.008115014 Vs 16:41:33 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = 0.0100000 m VT = -0.008165228 Vs 16:41:48 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = -0.0100000 m VT = 0.008123522 Vs 16:41:48 Integrated Voltage Average and RMS of Measurements X0 = -0.0200000 m, Delta X = 0.0100000 m VT = -0.008144411 +- 0.000002733 Vs 16:41:48 Integrated Field Strength, Average and RMS of Measurements X0 = -0.0200000 m, dx = 0.0100000 m, Nturns = 10.0 VT = -0.008144411 +- 0.000002733 Vs BL = -0.08144411 +- 0.00002733 Tm 16:41:48 Integrated Voltage Measurement... 16:42:06 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = 0.0100000 m VT = -0.004201288 Vs 16:42:21 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = -0.0100000 m VT = 0.004170279 Vs 16:42:36 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = 0.0100000 m VT = -0.004192078 Vs 16:42:51 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = -0.0100000 m VT = 0.004174395 Vs 16:43:07 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = 0.0100000 m VT = -0.004189756 Vs 16:43:22 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = -0.0100000 m VT = 0.004169309 Vs 16:43:37 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = 0.0100000 m VT = -0.004198550 Vs 16:43:52 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = -0.0100000 m VT = 0.004164951 Vs 16:43:52 Integrated Voltage Average and RMS of Measurements X0 = -0.0100000 m, Delta X = 0.0100000 m VT = -0.004182576 +- 0.000002273 Vs 16:43:52 Integrated Field Strength, Average and RMS of Measurements X0 = -0.0100000 m, dx = 0.0100000 m, Nturns = 10.0 VT = -0.004182576 +- 0.000002273 Vs BL = -0.04182576 +- 0.00002273 Tm 16:43:52 Integrated Voltage Measurement... 16:44:10 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = -0.000259204 Vs 16:44:25 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = 0.000195284 Vs 16:44:40 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = -0.000270324 Vs 16:44:55 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = 0.000196868 Vs 16:45:10 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = -0.000256347 Vs 16:45:25 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = 0.000197233 Vs 16:45:41 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = -0.000258766 Vs 16:45:56 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = 0.000199939 Vs 16:45:56 Integrated Voltage Average and RMS of Measurements X0 = 0.0000000 m, Delta X = 0.0100000 m VT = -0.000229246 +- 0.000002691 Vs 16:45:56 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0000000 m, dx = 0.0100000 m, Nturns = 10.0 VT = -0.000229246 +- 0.000002691 Vs BL = -0.00229246 +- 0.00002691 Tm 16:45:57 Integrated Voltage Measurement... 16:46:15 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = 0.0100000 m VT = 0.003698415 Vs 16:46:30 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = -0.0100000 m VT = -0.003760621 Vs 16:46:45 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = 0.0100000 m VT = 0.003694423 Vs 16:47:00 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = -0.0100000 m VT = -0.003755741 Vs 16:47:15 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = 0.0100000 m VT = 0.003697290 Vs 16:47:30 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = -0.0100000 m VT = -0.003754430 Vs 16:47:45 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = 0.0100000 m VT = 0.003698756 Vs 16:48:00 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = -0.0100000 m VT = -0.003747187 Vs 16:48:00 Integrated Voltage Average and RMS of Measurements X0 = 0.0100000 m, Delta X = 0.0100000 m VT = 0.003725858 +- 0.000002363 Vs 16:48:00 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0100000 m, dx = 0.0100000 m, Nturns = 10.0 VT = 0.003725858 +- 0.000002363 Vs BL = 0.03725858 +- 0.00002363 Tm 16:48:00 Integrated Voltage Measurement... 16:48:19 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = 0.0100000 m VT = 0.007669082 Vs 16:48:34 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = -0.0100000 m VT = -0.007696380 Vs 16:48:49 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = 0.0100000 m VT = 0.007675211 Vs 16:49:04 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = -0.0100000 m VT = -0.007683356 Vs 16:49:19 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = 0.0100000 m VT = 0.007666194 Vs 16:49:35 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = -0.0100000 m VT = -0.007695882 Vs 16:49:50 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = 0.0100000 m VT = 0.007673075 Vs 16:50:05 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = -0.0100000 m VT = -0.007698878 Vs 16:50:05 Integrated Voltage Average and RMS of Measurements X0 = 0.0200000 m, Delta X = 0.0100000 m VT = 0.007682257 +- 0.000002469 Vs 16:50:05 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0200000 m, dx = 0.0100000 m, Nturns = 10.0 VT = 0.007682257 +- 0.000002469 Vs BL = 0.07682257 +- 0.00002469 Tm 16:50:11 BL vs X Fit Coefficients: Cn sigCn n (Tm/m^n) ----- --------+-------- 0 -0.00230 0.00001 1 3.95618 0.00081 Compute Fit Residuals: X0 BL sigBL BLfit BL-BLfit (m) (Tm) (Tm) (Tm) (Tm) ---------- --------+-------- -------- -------- -0.020000 -0.08144 0.00003 -0.08142 -0.00002 -0.010000 -0.04183 0.00002 -0.04186 0.00003 0.000000 -0.00229 0.00003 -0.00230 0.00000 0.010000 0.03726 0.00002 0.03727 -0.00001 0.020000 0.07682 0.00002 0.07683 -0.00000 16:50:11 Magnet Strength Calculation Harmonic Number = 2 Number Of X0 Positions = 5 X0 BL sigBL (m) (Tm) (Tm) -------- --------+-------- -0.0200 -0.08144 0.00003 -0.0100 -0.04183 0.00002 0.0000 -0.00229 0.00003 0.0100 0.03726 0.00002 0.0200 0.07682 0.00002 SL = 3.9561770 +- 0.0008063 Tm^2/m^n 16:50:11 Quadrupole Strength Calculation Check Slope = 3.956177 Tm/m, Inter = -0.002296 Tm, Sig_slope = 0.000757 Tm/m, Est_sig = 0.000024 Tm GL = 3.956177 +- 0.000757 T 16:50:11 Magnet Strength Measurement Harmonic Number = 2 SL = 3.9561770 +- 0.0008063 Tm^2/m^n 16:50:12 Magnet Current, Imag = 30.1033 A Vtrans = 0.301033 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 16:50:13 Magnet Current, Imag = 30.1025 A Vtrans = 0.301025 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 16:50:15 Magnet Current, Imag = 30.1022 A Vtrans = 0.301022 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 16:50:16 Magnet Current, Imag = 30.1029 A Vtrans = 0.301029 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 16:50:16 Magnet Current Measurement: Imag = 30.1027 +- 0.0004 A 16:50:16 Magnet current ramping to 20.0000 A... 16:50:28 Magnet Current, Imag = 20.0560 A Vtrans = 0.200560 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 16:50:30 Magnet Current, Imag = 20.0557 A Vtrans = 0.200557 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 16:50:32 Magnet Current, Imag = 20.0554 A Vtrans = 0.200554 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 16:50:33 Magnet Current, Imag = 20.0559 A Vtrans = 0.200559 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 16:50:33 Magnet Current Measurement: Imag = 20.0558 +- 0.0002 A 16:50:33 Magnetic Field Measurement Measuring the magnet strength... 16:50:33 Integrated Voltage Measurement... 16:50:56 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = 0.0100000 m VT = -0.005467357 Vs 16:51:11 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = -0.0100000 m VT = 0.005419044 Vs 16:51:26 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = 0.0100000 m VT = -0.005473508 Vs 16:51:41 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = -0.0100000 m VT = 0.005425224 Vs 16:51:56 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = 0.0100000 m VT = -0.005477488 Vs 16:52:11 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = -0.0100000 m VT = 0.005417122 Vs 16:52:26 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = 0.0100000 m VT = -0.005482050 Vs 16:52:41 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = -0.0100000 m VT = 0.005410973 Vs 16:52:41 Integrated Voltage Average and RMS of Measurements X0 = -0.0200000 m, Delta X = 0.0100000 m VT = -0.005446596 +- 0.000002220 Vs 16:52:41 Integrated Field Strength, Average and RMS of Measurements X0 = -0.0200000 m, dx = 0.0100000 m, Nturns = 10.0 VT = -0.005446596 +- 0.000002220 Vs BL = -0.05446596 +- 0.00002220 Tm 16:52:41 Integrated Voltage Measurement... 16:53:00 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = 0.0100000 m VT = -0.002839025 Vs 16:53:15 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = -0.0100000 m VT = 0.002762629 Vs 16:53:30 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = 0.0100000 m VT = -0.002831578 Vs 16:53:45 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = -0.0100000 m VT = 0.002761617 Vs 16:54:00 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = 0.0100000 m VT = -0.002841263 Vs 16:54:15 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = -0.0100000 m VT = 0.002757323 Vs 16:54:30 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = 0.0100000 m VT = -0.002837913 Vs 16:54:45 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = -0.0100000 m VT = 0.002772472 Vs 16:54:45 Integrated Voltage Average and RMS of Measurements X0 = -0.0100000 m, Delta X = 0.0100000 m VT = -0.002800477 +- 0.000003115 Vs 16:54:45 Integrated Field Strength, Average and RMS of Measurements X0 = -0.0100000 m, dx = 0.0100000 m, Nturns = 10.0 VT = -0.002800477 +- 0.000003115 Vs BL = -0.02800477 +- 0.00003115 Tm 16:54:45 Integrated Voltage Measurement... 16:55:04 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = -0.000170949 Vs 16:55:19 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = 0.000141999 Vs 16:55:34 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = -0.000164918 Vs 16:55:49 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = 0.000130926 Vs 16:56:04 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = -0.000172056 Vs 16:56:20 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = 0.000130849 Vs 16:56:35 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = -0.000171532 Vs 16:56:50 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = 0.000142677 Vs 16:56:50 Integrated Voltage Average and RMS of Measurements X0 = 0.0000000 m, Delta X = 0.0100000 m VT = -0.000153238 +- 0.000003771 Vs 16:56:50 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0000000 m, dx = 0.0100000 m, Nturns = 10.0 VT = -0.000153238 +- 0.000003771 Vs BL = -0.00153238 +- 0.00003771 Tm 16:56:50 Integrated Voltage Measurement... 16:57:08 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = 0.0100000 m VT = 0.002479714 Vs 16:57:23 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = -0.0100000 m VT = -0.002495304 Vs 16:57:38 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = 0.0100000 m VT = 0.002477132 Vs 16:57:53 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = -0.0100000 m VT = -0.002504776 Vs 16:58:08 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = 0.0100000 m VT = 0.002484477 Vs 16:58:23 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = -0.0100000 m VT = -0.002491592 Vs 16:58:39 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = 0.0100000 m VT = 0.002486367 Vs 16:58:54 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = -0.0100000 m VT = -0.002488781 Vs 16:58:54 Integrated Voltage Average and RMS of Measurements X0 = 0.0100000 m, Delta X = 0.0100000 m VT = 0.002488518 +- 0.000001421 Vs 16:58:54 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0100000 m, dx = 0.0100000 m, Nturns = 10.0 VT = 0.002488518 +- 0.000001421 Vs BL = 0.02488518 +- 0.00001421 Tm 16:58:54 Integrated Voltage Measurement... 16:59:12 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = 0.0100000 m VT = 0.005130671 Vs 16:59:27 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = -0.0100000 m VT = -0.005124274 Vs 16:59:42 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = 0.0100000 m VT = 0.005145172 Vs 16:59:57 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = -0.0100000 m VT = -0.005118783 Vs 17:00:12 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = 0.0100000 m VT = 0.005139164 Vs 17:00:27 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = -0.0100000 m VT = -0.005112879 Vs 17:00:42 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = 0.0100000 m VT = 0.005138691 Vs 17:00:57 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = -0.0100000 m VT = -0.005114051 Vs 17:00:57 Integrated Voltage Average and RMS of Measurements X0 = 0.0200000 m, Delta X = 0.0100000 m VT = 0.005127961 +- 0.000002380 Vs 17:00:58 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0200000 m, dx = 0.0100000 m, Nturns = 10.0 VT = 0.005127961 +- 0.000002380 Vs BL = 0.05127961 +- 0.00002380 Tm 17:01:03 BL vs X Fit Coefficients: Cn sigCn n (Tm/m^n) ----- --------+-------- 0 -0.00157 0.00001 1 2.64381 0.00074 Compute Fit Residuals: X0 BL sigBL BLfit BL-BLfit (m) (Tm) (Tm) (Tm) (Tm) ---------- --------+-------- -------- -------- -0.020000 -0.05447 0.00002 -0.05444 -0.00002 -0.010000 -0.02800 0.00003 -0.02801 0.00000 0.000000 -0.00153 0.00004 -0.00157 0.00004 0.010000 0.02489 0.00001 0.02487 0.00001 0.020000 0.05128 0.00002 0.05131 -0.00003 17:01:03 Magnet Strength Calculation Harmonic Number = 2 Number Of X0 Positions = 5 X0 BL sigBL (m) (Tm) (Tm) -------- --------+-------- -0.0200 -0.05447 0.00002 -0.0100 -0.02800 0.00003 0.0000 -0.00153 0.00004 0.0100 0.02489 0.00001 0.0200 0.05128 0.00002 SL = 2.6438108 +- 0.0007355 Tm^2/m^n 17:01:03 Quadrupole Strength Calculation Check Slope = 2.643811 Tm/m, Inter = -0.001568 Tm, Sig_slope = 0.000964 Tm/m, Est_sig = 0.000030 Tm GL = 2.643811 +- 0.000964 T 17:01:03 Magnet Strength Measurement Harmonic Number = 2 SL = 2.6438108 +- 0.0007355 Tm^2/m^n 17:01:05 Magnet Current, Imag = 20.0552 A Vtrans = 0.200552 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 17:01:06 Magnet Current, Imag = 20.0553 A Vtrans = 0.200553 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 17:01:08 Magnet Current, Imag = 20.0562 A Vtrans = 0.200562 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 17:01:10 Magnet Current, Imag = 20.0551 A Vtrans = 0.200551 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 17:01:10 Magnet Current Measurement: Imag = 20.0554 +- 0.0005 A 17:01:10 Magnet current ramping to 10.0000 A... 17:01:23 Magnet Current, Imag = 10.0099 A Vtrans = 0.100099 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 17:01:24 Magnet Current, Imag = 10.0100 A Vtrans = 0.100100 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 17:01:26 Magnet Current, Imag = 10.0097 A Vtrans = 0.100097 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 17:01:28 Magnet Current, Imag = 10.0094 A Vtrans = 0.100094 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 17:01:28 Magnet Current Measurement: Imag = 10.0098 +- 0.0002 A 17:01:28 Magnetic Field Measurement Measuring the magnet strength... 17:01:28 Integrated Voltage Measurement... 17:01:50 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = 0.0100000 m VT = -0.002744576 Vs 17:02:05 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = -0.0100000 m VT = 0.002736005 Vs 17:02:20 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = 0.0100000 m VT = -0.002761153 Vs 17:02:35 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = -0.0100000 m VT = 0.002728107 Vs 17:02:51 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = 0.0100000 m VT = -0.002764096 Vs 17:03:06 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = -0.0100000 m VT = 0.002726046 Vs 17:03:21 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = 0.0100000 m VT = -0.002768691 Vs 17:03:37 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = -0.0100000 m VT = 0.002715057 Vs 17:03:37 Integrated Voltage Average and RMS of Measurements X0 = -0.0200000 m, Delta X = 0.0100000 m VT = -0.002742966 +- 0.000001972 Vs 17:03:37 Integrated Field Strength, Average and RMS of Measurements X0 = -0.0200000 m, dx = 0.0100000 m, Nturns = 10.0 VT = -0.002742966 +- 0.000001972 Vs BL = -0.02742966 +- 0.00001972 Tm 17:03:37 Integrated Voltage Measurement... 17:03:55 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = 0.0100000 m VT = -0.001439576 Vs 17:04:10 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = -0.0100000 m VT = 0.001375572 Vs 17:04:25 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = 0.0100000 m VT = -0.001455402 Vs 17:04:40 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = -0.0100000 m VT = 0.001369158 Vs 17:04:55 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = 0.0100000 m VT = -0.001453491 Vs 17:05:10 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = -0.0100000 m VT = 0.001364602 Vs 17:05:25 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = 0.0100000 m VT = -0.001448815 Vs 17:05:40 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = -0.0100000 m VT = 0.001391737 Vs 17:05:40 Integrated Voltage Average and RMS of Measurements X0 = -0.0100000 m, Delta X = 0.0100000 m VT = -0.001412294 +- 0.000004913 Vs 17:05:40 Integrated Field Strength, Average and RMS of Measurements X0 = -0.0100000 m, dx = 0.0100000 m, Nturns = 10.0 VT = -0.001412294 +- 0.000004913 Vs BL = -0.01412294 +- 0.00004913 Tm 17:05:41 Integrated Voltage Measurement... 17:05:59 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = -0.000110646 Vs 17:06:14 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = 0.000048508 Vs 17:06:29 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = -0.000126878 Vs 17:06:44 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = 0.000044605 Vs 17:06:59 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = -0.000128822 Vs 17:07:14 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = 0.000034823 Vs 17:07:29 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = -0.000115769 Vs 17:07:45 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = 0.000044458 Vs 17:07:45 Integrated Voltage Average and RMS of Measurements X0 = 0.0000000 m, Delta X = 0.0100000 m VT = -0.000081814 +- 0.000002415 Vs 17:07:45 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0000000 m, dx = 0.0100000 m, Nturns = 10.0 VT = -0.000081814 +- 0.000002415 Vs BL = -0.00081814 +- 0.00002415 Tm 17:07:45 Integrated Voltage Measurement... 17:08:03 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = 0.0100000 m VT = 0.001210177 Vs 17:08:18 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = -0.0100000 m VT = -0.001284993 Vs 17:08:33 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = 0.0100000 m VT = 0.001221590 Vs 17:08:48 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = -0.0100000 m VT = -0.001270849 Vs 17:09:03 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = 0.0100000 m VT = 0.001224236 Vs 17:09:18 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = -0.0100000 m VT = -0.001276552 Vs 17:09:33 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = 0.0100000 m VT = 0.001227685 Vs 17:09:48 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = -0.0100000 m VT = -0.001268427 Vs 17:09:49 Integrated Voltage Average and RMS of Measurements X0 = 0.0100000 m, Delta X = 0.0100000 m VT = 0.001248063 +- 0.000001505 Vs 17:09:49 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0100000 m, dx = 0.0100000 m, Nturns = 10.0 VT = 0.001248063 +- 0.000001505 Vs BL = 0.01248063 +- 0.00001505 Tm 17:09:49 Integrated Voltage Measurement... 17:10:07 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = 0.0100000 m VT = 0.002561257 Vs 17:10:22 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = -0.0100000 m VT = -0.002593828 Vs 17:10:37 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = 0.0100000 m VT = 0.002562882 Vs 17:10:52 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = -0.0100000 m VT = -0.002591720 Vs 17:11:07 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = 0.0100000 m VT = 0.002565228 Vs 17:11:22 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = -0.0100000 m VT = -0.002588893 Vs 17:11:38 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = 0.0100000 m VT = 0.002570361 Vs 17:11:53 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = -0.0100000 m VT = -0.002592252 Vs 17:11:53 Integrated Voltage Average and RMS of Measurements X0 = 0.0200000 m, Delta X = 0.0100000 m VT = 0.002578303 +- 0.000001743 Vs 17:11:53 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0200000 m, dx = 0.0100000 m, Nturns = 10.0 VT = 0.002578303 +- 0.000001743 Vs BL = 0.02578303 +- 0.00001743 Tm 17:11:58 BL vs X Fit Coefficients: Cn sigCn n (Tm/m^n) ----- --------+-------- 0 -0.00082 0.00001 1 1.33029 0.00074 Compute Fit Residuals: X0 BL sigBL BLfit BL-BLfit (m) (Tm) (Tm) (Tm) (Tm) ---------- --------+-------- -------- -------- -0.020000 -0.02743 0.00002 -0.02743 -0.00000 -0.010000 -0.01412 0.00005 -0.01412 0.00000 0.000000 -0.00082 0.00002 -0.00082 0.00000 0.010000 0.01248 0.00002 0.01248 -0.00000 0.020000 0.02578 0.00002 0.02578 -0.00000 17:11:58 Magnet Strength Calculation Harmonic Number = 2 Number Of X0 Positions = 5 X0 BL sigBL (m) (Tm) (Tm) -------- --------+-------- -0.0200 -0.02743 0.00002 -0.0100 -0.01412 0.00005 0.0000 -0.00082 0.00002 0.0100 0.01248 0.00002 0.0200 0.02578 0.00002 SL = 1.3302896 +- 0.0007355 Tm^2/m^n 17:11:58 Quadrupole Strength Calculation Check Slope = 1.330290 Tm/m, Inter = -0.000821 Tm, Sig_slope = 0.000084 Tm/m, Est_sig = 0.000003 Tm GL = 1.330290 +- 0.000084 T 17:11:58 Magnet Strength Measurement Harmonic Number = 2 SL = 1.3302896 +- 0.0007355 Tm^2/m^n 17:12:00 Magnet Current, Imag = 10.0084 A Vtrans = 0.100084 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 17:12:02 Magnet Current, Imag = 10.0088 A Vtrans = 0.100088 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 17:12:03 Magnet Current, Imag = 10.0084 A Vtrans = 0.100084 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 17:12:05 Magnet Current, Imag = 10.0090 A Vtrans = 0.100090 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 17:12:05 Magnet Current Measurement: Imag = 10.0086 +- 0.0003 A 17:12:06 Magnet current ramping to 0.0000 A...