SLAC Magnetic Measurements Date: 03-05-2004 Time: 11:55:51 Project: Orion Magnet Type: solenoid Magnet Name: Ecomp Serial Number: Run Number: 29 Measurement Device: B9900 Hall probe Operator: keh Comment: post fix, not aligned yet Standardization: 4 cycles from 0.0 A to 220.0 A to 0.0 A Test Currents (A): 150.0 11:56:10 Stage Zeroed 11:56:10 MoveZ Devices: Move Device = SXDRIVE, Read Device = SXDRIVE 11:57:08 Move Stage, position = -0.250000 11:58:23 Hall Probe Zeroed 11:58:45 Hall Probe Type Being Used, type = B9900 11:59:46 Move Stage, position = 0.000000 11:59:47 Initial Zero Current Measurement, Imag_zero = 0.0000 A 11:59:51 Hall Probe Range Set, range = 4 11:59:51 Magnet current ramping to 220.0000 A... 12:02:27 Magnet current ramping to 0.0000 A... 12:05:02 Magnet current ramping to 220.0000 A... 12:07:39 Magnet current ramping to 0.0000 A... 12:10:15 Magnet current ramping to 220.0000 A... 12:12:51 Magnet current ramping to 0.0000 A... 12:15:27 Magnet current ramping to 220.0000 A... 12:18:03 Magnet current ramping to 0.0000 A... 12:20:39 Magnet current ramping to 150.0000 A... 12:23:31 Move Stage, position = -0.250000 12:23:33 Stage Position Measurement, pos = -0.250000 m 12:23:34 Magnet Current, Imag = 150.0658 A Vtrans = 1.500658 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 12:23:34 Hall Probe Field Measurement, B = -0.000480 T 12:23:41 Move Stage, position = -0.240000 12:23:43 Stage Position Measurement, pos = -0.240000 m 12:23:44 Magnet Current, Imag = 150.0642 A Vtrans = 1.500642 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 12:23:45 Hall Probe Field Measurement, B = -0.000570 T 12:23:52 Move Stage, position = -0.230000 12:23:53 Stage Position Measurement, pos = -0.230000 m 12:23:55 Magnet Current, Imag = 150.0678 A Vtrans = 1.500678 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 12:23:55 Hall Probe Field Measurement, B = -0.000680 T 12:24:02 Move Stage, position = -0.220000 12:24:04 Stage Position Measurement, pos = -0.220000 m 12:24:05 Magnet Current, Imag = 150.0603 A Vtrans = 1.500603 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 12:24:05 Hall Probe Field Measurement, B = -0.000830 T 12:24:12 Move Stage, position = -0.210000 12:24:14 Stage Position Measurement, pos = -0.210000 m 12:24:15 Magnet Current, Imag = 150.0662 A Vtrans = 1.500662 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 12:24:16 Hall Probe Field Measurement, B = -0.001020 T 12:24:23 Move Stage, position = -0.200000 12:24:24 Stage Position Measurement, pos = -0.200000 m 12:24:26 Magnet Current, Imag = 150.0606 A Vtrans = 1.500606 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 12:24:26 Hall Probe Field Measurement, B = -0.001320 T 12:24:33 Move Stage, position = -0.190000 12:24:35 Stage Position Measurement, pos = -0.190000 m 12:24:36 Magnet Current, Imag = 150.0628 A Vtrans = 1.500628 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 12:24:37 Hall Probe Field Measurement, B = -0.001760 T 12:24:44 Move Stage, position = -0.180000 12:24:45 Stage Position Measurement, pos = -0.180000 m 12:24:46 Magnet Current, Imag = 150.0595 A Vtrans = 1.500595 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 12:24:47 Hall Probe Field Measurement, B = -0.002400 T 12:24:54 Move Stage, position = -0.170000 12:24:56 Stage Position Measurement, pos = -0.170000 m 12:24:57 Magnet Current, Imag = 150.0636 A Vtrans = 1.500636 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 12:24:57 Hall Probe Field Measurement, B = -0.003450 T 12:25:04 Move Stage, position = -0.160000 12:25:06 Stage Position Measurement, pos = -0.160000 m 12:25:07 Magnet Current, Imag = 150.0615 A Vtrans = 1.500615 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 12:25:08 Hall Probe Field Measurement, B = -0.005180 T 12:25:15 Move Stage, position = -0.150000 12:25:16 Stage Position Measurement, pos = -0.150000 m 12:25:18 Magnet Current, Imag = 150.0637 A Vtrans = 1.500637 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 12:25:18 Hall Probe Field Measurement, B = -0.008170 T 12:25:26 Move Stage, position = -0.140000 12:25:27 Stage Position Measurement, pos = -0.140000 m 12:25:28 Magnet Current, Imag = 150.0663 A Vtrans = 1.500663 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 12:25:29 Hall Probe Field Measurement, B = -0.013540 T 12:25:36 Move Stage, position = -0.130000 12:25:38 Stage Position Measurement, pos = -0.130000 m 12:25:39 Magnet Current, Imag = 150.0611 A Vtrans = 1.500611 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 12:25:39 Hall Probe Field Measurement, B = -0.023160 T 12:25:46 Move Stage, position = -0.120000 12:25:48 Stage Position Measurement, pos = -0.120000 m 12:25:49 Magnet Current, Imag = 150.0607 A Vtrans = 1.500607 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 12:25:50 Hall Probe Field Measurement, B = -0.040060 T 12:25:57 Move Stage, position = -0.110000 12:25:58 Stage Position Measurement, pos = -0.110000 m 12:25:59 Magnet Current, Imag = 150.0616 A Vtrans = 1.500616 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 12:26:00 Hall Probe Field Measurement, B = -0.066400 T 12:26:07 Move Stage, position = -0.100000 12:26:09 Stage Position Measurement, pos = -0.100000 m 12:26:10 Magnet Current, Imag = 150.0606 A Vtrans = 1.500606 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 12:26:10 Hall Probe Field Measurement, B = -0.101580 T 12:26:17 Move Stage, position = -0.090000 12:26:19 Stage Position Measurement, pos = -0.090000 m 12:26:20 Magnet Current, Imag = 150.0611 A Vtrans = 1.500611 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 12:26:21 Hall Probe Field Measurement, B = -0.138740 T 12:26:28 Move Stage, position = -0.080000 12:26:29 Stage Position Measurement, pos = -0.080000 m 12:26:31 Magnet Current, Imag = 150.0648 A Vtrans = 1.500648 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 12:26:31 Hall Probe Field Measurement, B = -0.169800 T 12:26:38 Move Stage, position = -0.070000 12:26:40 Stage Position Measurement, pos = -0.070000 m 12:26:41 Magnet Current, Imag = 150.0655 A Vtrans = 1.500655 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 12:26:41 Hall Probe Field Measurement, B = -0.191730 T 12:26:49 Move Stage, position = -0.060000 12:26:50 Stage Position Measurement, pos = -0.060000 m 12:26:51 Magnet Current, Imag = 150.0630 A Vtrans = 1.500630 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 12:26:52 Hall Probe Field Measurement, B = -0.205560 T 12:26:59 Move Stage, position = -0.050000 12:27:01 Stage Position Measurement, pos = -0.050000 m 12:27:02 Magnet Current, Imag = 150.0641 A Vtrans = 1.500641 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 12:27:02 Hall Probe Field Measurement, B = -0.213760 T 12:27:09 Move Stage, position = -0.040000 12:27:11 Stage Position Measurement, pos = -0.040000 m 12:27:12 Magnet Current, Imag = 150.0592 A Vtrans = 1.500592 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 12:27:13 Hall Probe Field Measurement, B = -0.218450 T 12:27:20 Move Stage, position = -0.030000 12:27:21 Stage Position Measurement, pos = -0.030000 m 12:27:23 Magnet Current, Imag = 150.0606 A Vtrans = 1.500606 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 12:27:23 Hall Probe Field Measurement, B = -0.221090 T 12:27:30 Move Stage, position = -0.020000 12:27:32 Stage Position Measurement, pos = -0.020000 m 12:27:33 Magnet Current, Imag = 150.0637 A Vtrans = 1.500637 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 12:27:33 Hall Probe Field Measurement, B = -0.222500 T 12:27:40 Move Stage, position = -0.010000 12:27:42 Stage Position Measurement, pos = -0.010000 m 12:27:43 Magnet Current, Imag = 150.0624 A Vtrans = 1.500624 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 12:27:44 Hall Probe Field Measurement, B = -0.223100 T 12:27:51 Move Stage, position = 0.000000 12:27:52 Stage Position Measurement, pos = 0.000000 m 12:27:54 Magnet Current, Imag = 150.0637 A Vtrans = 1.500637 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 12:27:54 Hall Probe Field Measurement, B = -0.223050 T 12:28:01 Move Stage, position = 0.010000 12:28:03 Stage Position Measurement, pos = 0.010000 m 12:28:04 Magnet Current, Imag = 150.0675 A Vtrans = 1.500675 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 12:28:05 Hall Probe Field Measurement, B = -0.222550 T 12:28:12 Move Stage, position = 0.020000 12:28:13 Stage Position Measurement, pos = 0.020000 m 12:28:14 Magnet Current, Imag = 150.0665 A Vtrans = 1.500665 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 12:28:15 Hall Probe Field Measurement, B = -0.221510 T 12:28:22 Move Stage, position = 0.030000 12:28:24 Stage Position Measurement, pos = 0.030000 m 12:28:25 Magnet Current, Imag = 150.0629 A Vtrans = 1.500629 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 12:28:25 Hall Probe Field Measurement, B = -0.219790 T 12:28:32 Move Stage, position = 0.040000 12:28:34 Stage Position Measurement, pos = 0.040000 m 12:28:35 Magnet Current, Imag = 150.0638 A Vtrans = 1.500638 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 12:28:36 Hall Probe Field Measurement, B = -0.216840 T 12:28:43 Move Stage, position = 0.050000 12:28:44 Stage Position Measurement, pos = 0.050000 m 12:28:46 Magnet Current, Imag = 150.0640 A Vtrans = 1.500640 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 12:28:46 Hall Probe Field Measurement, B = -0.211790 T 12:28:53 Move Stage, position = 0.060000 12:28:55 Stage Position Measurement, pos = 0.060000 m 12:28:56 Magnet Current, Imag = 150.0603 A Vtrans = 1.500603 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 12:28:57 Hall Probe Field Measurement, B = -0.203070 T 12:29:04 Move Stage, position = 0.070000 12:29:05 Stage Position Measurement, pos = 0.070000 m 12:29:06 Magnet Current, Imag = 150.0604 A Vtrans = 1.500604 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 12:29:07 Hall Probe Field Measurement, B = -0.188660 T 12:29:14 Move Stage, position = 0.080000 12:29:16 Stage Position Measurement, pos = 0.080000 m 12:29:17 Magnet Current, Imag = 150.0554 A Vtrans = 1.500554 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 12:29:17 Hall Probe Field Measurement, B = -0.166110 T 12:29:24 Move Stage, position = 0.090000 12:29:26 Stage Position Measurement, pos = 0.090000 m 12:29:27 Magnet Current, Imag = 150.0625 A Vtrans = 1.500625 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 12:29:28 Hall Probe Field Measurement, B = -0.134560 T 12:29:35 Move Stage, position = 0.100000 12:29:36 Stage Position Measurement, pos = 0.100000 m 12:29:38 Magnet Current, Imag = 150.0612 A Vtrans = 1.500612 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 12:29:38 Hall Probe Field Measurement, B = -0.097540 T 12:29:45 Move Stage, position = 0.110000 12:29:47 Stage Position Measurement, pos = 0.110000 m 12:29:48 Magnet Current, Imag = 150.0561 A Vtrans = 1.500561 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 12:29:48 Hall Probe Field Measurement, B = -0.063240 T 12:29:56 Move Stage, position = 0.120000 12:29:57 Stage Position Measurement, pos = 0.120000 m 12:29:58 Magnet Current, Imag = 150.0617 A Vtrans = 1.500617 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 12:29:59 Hall Probe Field Measurement, B = -0.037910 T 12:30:06 Move Stage, position = 0.130000 12:30:08 Stage Position Measurement, pos = 0.130000 m 12:30:09 Magnet Current, Imag = 150.0633 A Vtrans = 1.500633 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 12:30:10 Hall Probe Field Measurement, B = -0.021900 T 12:30:17 Move Stage, position = 0.140000 12:30:18 Stage Position Measurement, pos = 0.140000 m 12:30:20 Magnet Current, Imag = 150.0669 A Vtrans = 1.500669 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 12:30:20 Hall Probe Field Measurement, B = -0.012830 T 12:30:27 Move Stage, position = 0.150000 12:30:29 Stage Position Measurement, pos = 0.150000 m 12:30:30 Magnet Current, Imag = 150.0602 A Vtrans = 1.500602 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 12:30:30 Hall Probe Field Measurement, B = -0.007760 T 12:30:38 Move Stage, position = 0.160000 12:30:39 Stage Position Measurement, pos = 0.160000 m 12:30:40 Magnet Current, Imag = 150.0610 A Vtrans = 1.500610 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 12:30:41 Hall Probe Field Measurement, B = -0.004940 T 12:30:48 Move Stage, position = 0.170000 12:30:50 Stage Position Measurement, pos = 0.170000 m 12:30:51 Magnet Current, Imag = 150.0569 A Vtrans = 1.500569 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 12:30:51 Hall Probe Field Measurement, B = -0.003300 T 12:30:58 Move Stage, position = 0.180000 12:31:00 Stage Position Measurement, pos = 0.180000 m 12:31:01 Magnet Current, Imag = 150.0631 A Vtrans = 1.500631 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 12:31:02 Hall Probe Field Measurement, B = -0.002320 T 12:31:09 Move Stage, position = 0.190000 12:31:10 Stage Position Measurement, pos = 0.190000 m 12:31:11 Magnet Current, Imag = 150.0557 A Vtrans = 1.500557 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 12:31:12 Hall Probe Field Measurement, B = -0.001680 T 12:31:19 Move Stage, position = 0.200000 12:31:21 Stage Position Measurement, pos = 0.200000 m 12:31:22 Magnet Current, Imag = 150.0579 A Vtrans = 1.500579 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 12:31:22 Hall Probe Field Measurement, B = -0.001270 T 12:31:29 Move Stage, position = 0.210000 12:31:31 Stage Position Measurement, pos = 0.210000 m 12:31:32 Magnet Current, Imag = 150.0647 A Vtrans = 1.500647 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 12:31:33 Hall Probe Field Measurement, B = -0.001000 T 12:31:40 Move Stage, position = 0.220000 12:31:41 Stage Position Measurement, pos = 0.220000 m 12:31:43 Magnet Current, Imag = 150.0620 A Vtrans = 1.500620 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 12:31:43 Hall Probe Field Measurement, B = -0.000800 T 12:31:50 Move Stage, position = 0.230000 12:31:52 Stage Position Measurement, pos = 0.230000 m 12:31:53 Magnet Current, Imag = 150.0559 A Vtrans = 1.500559 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 12:31:53 Hall Probe Field Measurement, B = -0.000650 T 12:32:01 Move Stage, position = 0.240000 12:32:02 Stage Position Measurement, pos = 0.240000 m 12:32:03 Magnet Current, Imag = 150.0576 A Vtrans = 1.500576 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 12:32:04 Hall Probe Field Measurement, B = -0.000540 T 12:32:11 Move Stage, position = 0.250000 12:32:13 Stage Position Measurement, pos = 0.250000 m 12:32:14 Magnet Current, Imag = 150.0622 A Vtrans = 1.500622 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 12:32:14 Hall Probe Field Measurement, B = -0.000470 T 12:33:09 Move Stage, position = 0.000000 12:33:10 Magnet current ramping to 0.0000 A...