SLAC Magnetic Measurements Date: 04-18-2008 Time: 14:00:27 Planar Insertion Device Analysis -------------------------------- Data File Header: SLAC Magnetic Measurements Date: 01-14-2008 Time: 09:42:36 Project: LCLS Device Type: Undulator Data Type: Serial Number: L143-112000-37 DATASET Number: 2 Tuning Stage: Fine Tuning Run Number: 3 Measurement Type: Z Scans Operator: YL Comment: find Y at min B Measurement Device x: Measurement Device y: SENTRON_441 Data File: s:\magdata\LCLS\Undulator\L143-112000-37\DATASET0002\Fine Tuning\Z Scans\Run 003\bscanz.dat Temperature (C) L143-112000-37:--> top_end 19.87 --> top_mid 19.88 --> top_start 19.90 --> bottom 19.88 --> Air 19.93 --> Granite 19.87 Axes position (m) X:0.030316 Y:0.000269 Initial Hall Probe Zero Measurements: Vx_0i = 0.001712 V Vy_0i = -0.000724 V Final Hall Probe Zero Measurements: Vx_0f = 0.001671 V Vy_0f = -0.000737 V Offset Correction Voff = offset+index*increment index = 0,1...21250 Vx:offset = 0.001712 V, increment = -1.965E-09 V Vy:offset = -0.000724 V, increment = -6.188E-10 V Hall Probe Calibration File: Y: c:\CVI\current_calib\Sentron_441\Sentron_441_y.ini X: c:\cvi\current_calib\HALLx_calib_poly_fit_file.dat Reference Magnet Measurements: Bx_ref = 0.0000xx (T) By_ref = 0.0000xx (T) Magnetic Field Measurements vs Z Analysis Parameters: Input file directory = S:\MagData\LCLS\Undulator\L143-112000-37\DATASET0002\Fine Tuning\Z Scans\Run 003\ Input file name = x+00000_y+000_bscanz.dat Input file type = ZVXYBXYH Correction required = 1 Correction, Bx multiplicative factor = 1.000000 Correction, By multiplicative factor = 1.000000 Correction, Bx additive constant = -0.000018 T Correction, By additive constant = 0.000007 T Correction Bxy additive file = none Correction Bxy additive file type = CORR_ZBXY Electron gamma = 26656.830000 Electron energy = 13.621615 Gev Nominal undulator period = 0.030000 m Nominal K value = 3.492230 Nominal radiation wavelength = 1.498305e-010 m Number of undulator poles = 226 Number of end poles (used in analysis) = 4 Cell length = 3.656000 m Nominal phase change through cell = 709.999940 rad Nominal phase of pole after end poles = 14.137167 rad Save as final result = 0 Result file directory = c:\magdata\slac_und_2\ Result file name = zscan_xn0060yp0002.dat Figures close on exit = 0 Figures visible = 1 !!!!!!!!!!!!!!!!!!!!!!!!!!!!!! The Measured Fields Were Corrected: Bx_corr = CORR_BX_MULT * Bx + CORR_BX_ADD + CORR_BX_ADD_FILE By_corr = CORR_BY_MULT * By + CORR_BY_ADD + CORR_BY_ADD_FILE Parameters: CORR_BX_MULT = 1.000000 CORR_BY_MULT = 1.000000 CORR_BX_ADD = -0.000018 T CORR_BY_ADD = 0.000007 T CORR_BX_ADD_FILE = none CORR_BY_ADD_FILE = none !!!!!!!!!!!!!!!!!!!!!!!!!!!!!! Peak And Zero Field Summary: Number of Poles = 226 Number of End Poles (excluded in statistics) = 4 Peak Fields: Ave = 1.273311 T, RMS = 0.004778 T Peak Field Spacing: Ave = 0.015001 m, RMS = 0.000022 m Zero Field Spacing: Ave = 0.015001 m, RMS = 0.000050 m Average Undulator Period = 0.030001 m Hall Probe Overall Field Integrals: Int Bx dz = -3.8361e-005 Tm Int Int Bx dz dz = -4.4577e-005 Tm^2 Int By dz = 5.3822e-005 Tm Int Int By dz dz = 9.8490e-005 Tm^2 Trajectory Exit Information: X at exit = 2.1676e-006 m X slope at exit = 1.1845e-006 Y at exit = 9.8107e-007 m Y slope at exit = 8.4426e-007 Calculated Undulator Parameters: Keff = 3.492412 Beff = 1.246707 T Lambda_rad = 1.498505e-010 m Calculated K Value At 20 C: (Assumes Keff changes by -.05% per degree C) Measured Average Undulator Temperature = 19.882500 C Measured Keff = 3.492412 Calculated Keff at 20 C = 3.492207 Relative Change (Keff_20C - Keff) / Keff = -0.000059 Phase Fluctuations: Slippage vs Z At Peaks Fitted Slope = 4.99480e-009 Undulator Period = 0.030001 m Radiation Wavelength = 1.49850e-010 m Phase Error (From Slippage vs Z Fit Residuals): Ave = 0.000000 deg, RMS = 4.514843 deg Phase Relative To Nominal Radiation Wave: Nominal Radiation Wavelength = 1.49830e-010 m Cell Length = 3.656000 m Number Of End Poles = 4 Entrance: Phase Error Of First 10 Poles After The End Poles Ave = -0.077521 deg, RMS = 3.552553 deg Center: Phase Error Of All Poles In Undulator Center: Ave = 3.546154 deg, RMS = 4.776743 deg Exit: Phase Error To Cell Boundary Of Last 10 Poles Before The End Poles Ave = 0.388662 deg, RMS = 4.253034 deg Cell: Phase Change Of Cell = 40685.742 deg Nominal Phase Change Of Cell = 40680.000 deg Phase Error Of Cell = 5.742 deg Calculated Phase Change In The Cell At 20 C: (Relative to the nominal radiation wave) (Assumes Keff changes by -.05% per degree C) Measured Average Undulator Temperature = 19.882500 C Measured Phase Change Of Cell = 40685.742 deg Calculated Phase Change Of Cell At 20 C = 40681.635 deg Phase Change Due To Temperature (Phase_20C - Phase) = -4.107 deg