SLAC Magnetic Measurements Date: 11-06-2007 Time: 15:27:33 Planar Insertion Device Analysis -------------------------------- Data File Header: SLAC Magnetic Measurements Date: 11-06-2007 Time: 15:19:38 Project: LCLS Device Type: Undulator Data Type: Serial Number: L143-112000-35 DATASET Number: 1 Tuning Stage: Fine Tuning Run Number: 7 Measurement Type: Z Scans Operator: YL Comment: phase shim #185 Measurement Device x: Measurement Device y: SENTRON_441 Data File: s:\magdata\LCLS\Undulator\L143-112000-35\DATASET0001\Fine Tuning\Z Scans\Run 007\bscanz.dat Temperature (C) L143-112000-35:--> top_end 19.91 --> top_mid 19.93 --> top_start 20.00 --> bottom 19.93 --> Air 20.02 --> Granite 19.92 Axes position (m) X:0.030183 Y:0.000257 Initial Hall Probe Zero Measurements: Vx_0i = 0.001462 V Vy_0i = -0.000897 V Final Hall Probe Zero Measurements: Vx_0f = 0.001508 V Vy_0f = -0.000924 V Offset Correction Voff = offset+index*increment index = 0,1...21250 Vx:offset = 0.001462 V, increment = 2.183E-09 V Vy:offset = -0.000897 V, increment = -1.273E-09 V Hall Probe Calibration File: Y: c:\CVI\current_calib\Sentron_441\Sentron_441_y.ini X: c:\cvi\current_calib\HALLx_calib_poly_fit_file.dat Reference Magnet Measurements: Bx_ref = 0.0000xx (T) By_ref = 0.0000xx (T) Magnetic Field Measurements vs Z Analysis Parameters: Input file directory = S:\MagData\LCLS\Undulator\L143-112000-35\DATASET0001\Fine Tuning\Z Scans\Run 007\ Input file name = x+00000_y+000_bscanz.dat Input file type = ZVXYBXYH Correction required = 1 Correction, Bx multiplicative factor = 1.000000 Correction, By multiplicative factor = 1.000000 Correction, Bx additive constant = -0.000006 T Correction, By additive constant = 0.000005 T Correction Bxy additive file = none Correction Bxy additive file type = CORR_ZBXY Electron gamma = 26652.520000 Electron energy = 13.619413 Gev Nominal undulator period = 0.030000 m Nominal K value = 3.491296 Nominal radiation wavelength = 1.498101e-010 m Number of undulator poles = 226 Number of end poles (used in analysis) = 4 Cell length = 3.656000 m Nominal phase change through cell = 709.999940 rad Nominal phase of pole after end poles = 14.137167 rad Save as final result = 0 Result file directory = c:\magdata\slac_und_2\ Result file name = zscan_xn0060yp0002.dat Figures close on exit = 0 Figures visible = 1 !!!!!!!!!!!!!!!!!!!!!!!!!!!!!! The Measured Fields Were Corrected: Bx_corr = CORR_BX_MULT * Bx + CORR_BX_ADD + CORR_BX_ADD_FILE By_corr = CORR_BY_MULT * By + CORR_BY_ADD + CORR_BY_ADD_FILE Parameters: CORR_BX_MULT = 1.000000 CORR_BY_MULT = 1.000000 CORR_BX_ADD = -0.000006 T CORR_BY_ADD = 0.000005 T CORR_BX_ADD_FILE = none CORR_BY_ADD_FILE = none !!!!!!!!!!!!!!!!!!!!!!!!!!!!!! Peak And Zero Field Summary: Number of Poles = 226 Number of End Poles (excluded in statistics) = 4 Peak Fields: Ave = 1.272435 T, RMS = 0.004993 T Peak Field Spacing: Ave = 0.015002 m, RMS = 0.000031 m Zero Field Spacing: Ave = 0.015002 m, RMS = 0.000040 m Average Undulator Period = 0.030004 m Hall Probe Overall Field Integrals: Int Bx dz = -7.3080e-006 Tm Int Int Bx dz dz = 2.7587e-005 Tm^2 Int By dz = -2.3710e-006 Tm Int Int By dz dz = -1.1520e-005 Tm^2 Trajectory Exit Information: X at exit = -2.5358e-007 m X slope at exit = -5.2191e-008 Y at exit = -6.0725e-007 m Y slope at exit = 1.6086e-007 Calculated Undulator Parameters: Keff = 3.491104 Beff = 1.246129 T Lambda_rad = 1.498159e-010 m Calculated K Value At 20 C: (Assumes Keff changes by -.05% per degree C) Measured Average Undulator Temperature = 19.942500 C Measured Keff = 3.491104 Calculated Keff at 20 C = 3.491004 Relative Change (Keff_20C - Keff) / Keff = -0.000029 Phase Fluctuations: Slippage vs Z At Peaks Fitted Slope = 4.99320e-009 Undulator Period = 0.030004 m Radiation Wavelength = 1.49816e-010 m Phase Error (From Slippage vs Z Fit Residuals): Ave = -0.000000 deg, RMS = 4.279883 deg Phase Relative To Nominal Radiation Wave: Nominal Radiation Wavelength = 1.49810e-010 m Cell Length = 3.656000 m Number Of End Poles = 4 Entrance: Phase Error Of First 10 Poles After The End Poles Ave = -0.009374 deg, RMS = 4.406115 deg Center: Phase Error Of All Poles In Undulator Center: Ave = -1.285024 deg, RMS = 4.272428 deg Exit: Phase Error To Cell Boundary Of Last 10 Poles Before The End Poles Ave = 1.025464 deg, RMS = 6.168960 deg Cell: Phase Change Of Cell = 40679.599 deg Nominal Phase Change Of Cell = 40680.000 deg Phase Error Of Cell = -0.401 deg Calculated Phase Change In The Cell At 20 C: (Relative to the nominal radiation wave) (Assumes Keff changes by -.05% per degree C) Measured Average Undulator Temperature = 19.942500 C Measured Phase Change Of Cell = 40679.599 deg Calculated Phase Change Of Cell At 20 C = 40677.589 deg Phase Change Due To Temperature (Phase_20C - Phase) = -2.009 deg