SLAC Magnetic Measurements Date: 03-07-2008 Time: 15:49:17 Planar Insertion Device Analysis -------------------------------- Data File Header: SLAC Magnetic Measurements Date: 03-07-2008 Time: 15:42:29 Project: LCLS Device Type: Undulator Data Type: Serial Number: L143-112000-30 DATASET Number: 1 Tuning Stage: Rough Tuning Run Number: 60 Measurement Type: Z Scans Operator: aw Comment: 1st pole = 0.370300, x=0.026158 y=0.00025607. 2 turns in at 38, 2 turns in at 51. Measurement Device x: Measurement Device y: SENIS_24-06 Data File: s:\magdata\LCLS\Undulator\L143-112000-30\DATASET0001\Rough Tuning\Z Scans\Run 060\bscanz.dat Temperature (C) L143-112000-30:--> top_start 20.16 --> top_mid 20.13 --> top_end 20.02 --> bot 20.11 --> granite 20.05 --> Air 19.99 Axes position (m) X:0.026158 Y:0.000256 Initial Hall Probe Zero Measurements: Vx_0i = -0.000352 V Vy_0i = -0.000449 V Final Hall Probe Zero Measurements: Vx_0f = -0.000358 V Vy_0f = -0.000475 V Offset Correction Voff = offset+index*increment index = 0,1...19850 Vx:offset = -0.000352 V, increment = -2.933E-10 V Vy:offset = -0.000449 V, increment = -1.330E-09 V Hall Probe Calibration File: Y: C:\CVI\current_calib\SENIS_24-06\SENIS_24-06_y.ini X: c:\cvi\current_calib\HALLx_calib_poly_fit_file.dat Reference Magnet Measurements: Bx_ref = 0.0000xx (T) By_ref = 0.0000xx (T) Magnetic Field Measurements vs Z Analysis Parameters: Input file directory = S:\MagData\LCLS\Undulator\L143-112000-30\DATASET0001\Rough Tuning\Z Scans\Run 060\ Input file name = x+00000_y+000_bscanz.dat Input file type = ZVXYBXYH Correction required = 1 Correction, Bx multiplicative factor = 1.000000 Correction, By multiplicative factor = 1.000000 Correction, Bx additive constant = 0.000015 T Correction, By additive constant = 0.000000 T Correction Bxy additive file = S:\Magdata\Earth Field Correction\earth_field_corr.dat Correction Bxy additive file type = CORR_ZBXY Electron gamma = 26682.250000 Electron energy = 13.634605 Gev Nominal undulator period = 0.030000 m Nominal K value = 3.497737 Nominal radiation wavelength = 1.499506e-010 m Number of undulator poles = 226 Number of end poles (used in analysis) = 4 Cell length = 3.656000 m Nominal phase change through cell = 709.999940 rad Nominal phase of pole after end poles = 14.137167 rad Save as final result = 0 Result file directory = c:\magdata\slac_und_2\ Result file name = zscan_xn0060yp0002.dat Figures close on exit = 0 Figures visible = 1 !!!!!!!!!!!!!!!!!!!!!!!!!!!!!! The Measured Fields Were Corrected: Bx_corr = CORR_BX_MULT * Bx + CORR_BX_ADD + CORR_BX_ADD_FILE By_corr = CORR_BY_MULT * By + CORR_BY_ADD + CORR_BY_ADD_FILE Parameters: CORR_BX_MULT = 1.000000 CORR_BY_MULT = 1.000000 CORR_BX_ADD = 0.000015 T CORR_BY_ADD = 0.000000 T CORR_BX_ADD_FILE = S:\Magdata\Earth Field Correction\earth_field_corr.dat CORR_BY_ADD_FILE = S:\Magdata\Earth Field Correction\earth_field_corr.dat !!!!!!!!!!!!!!!!!!!!!!!!!!!!!! Peak And Zero Field Summary: Number of Poles = 226 Number of End Poles (excluded in statistics) = 4 Peak Fields: Ave = 1.275623 T, RMS = 0.006953 T Peak Field Spacing: Ave = 0.015001 m, RMS = 0.000028 m Zero Field Spacing: Ave = 0.015001 m, RMS = 0.000048 m Average Undulator Period = 0.030002 m Hall Probe Overall Field Integrals: Int Bx dz = 5.0415e-005 Tm Int Int Bx dz dz = 5.3916e-005 Tm^2 Int By dz = -1.1106e-004 Tm Int Int By dz dz = -2.7332e-004 Tm^2 Trajectory Exit Information: X at exit = -6.0097e-006 m X slope at exit = -2.4418e-006 Y at exit = -1.1855e-006 m Y slope at exit = -1.1085e-006 Calculated Undulator Parameters: Keff = 3.498204 Beff = 1.248764 T Lambda_rad = 1.499929e-010 m Calculated K Value At 20 C: (Assumes Keff changes by -.05% per degree C) Measured Average Undulator Temperature = 20.105000 C Measured Keff = 3.498204 Calculated Keff at 20 C = 3.498387 Relative Change (Keff_20C - Keff) / Keff = 0.000052 Phase Fluctuations: Slippage vs Z At Peaks Fitted Slope = 4.99950e-009 Undulator Period = 0.030002 m Radiation Wavelength = 1.49993e-010 m Phase Error (From Slippage vs Z Fit Residuals): Ave = 0.000000 deg, RMS = 5.276578 deg Phase Relative To Nominal Radiation Wave: Nominal Radiation Wavelength = 1.49951e-010 m Cell Length = 3.656000 m Number Of End Poles = 4 Entrance: Phase Error Of First 10 Poles After The End Poles Ave = 0.055573 deg, RMS = 5.984075 deg Center: Phase Error Of All Poles In Undulator Center: Ave = -0.206899 deg, RMS = 5.987923 deg Exit: Phase Error To Cell Boundary Of Last 10 Poles Before The End Poles Ave = 4.279395 deg, RMS = 6.703146 deg Cell: Phase Change Of Cell = 40690.689 deg Nominal Phase Change Of Cell = 40680.000 deg Phase Error Of Cell = 10.689 deg Calculated Phase Change In The Cell At 20 C: (Relative to the nominal radiation wave) (Assumes Keff changes by -.05% per degree C) Measured Average Undulator Temperature = 20.105000 C Measured Phase Change Of Cell = 40690.689 deg Calculated Phase Change Of Cell At 20 C = 40694.361 deg Phase Change Due To Temperature (Phase_20C - Phase) = 3.672 deg