SLAC Magnetic Measurements
Date: 03-13-2007
Time: 12:28:41
 
Planar Insertion Device Analysis
--------------------------------

Data File Header:
SLAC Magnetic Measurements
Date: 03-13-2007
Time: 12:12:43
Project: LCLS
Device Type: Undulator
Data Type: 
Serial Number: L143-112000-17
DATASET Number: 1
Tuning Stage: Rough Tuning
Run Number: 146
Measurement Type: Z Scans
Operator: AW
Comment: y=0.000089(for By)  At x-shim at 15, turned in screw by 2 1/2 turns.
Measurement Device x: 
Measurement Device y: SENIS_24-06
Data File: s:\magdata\LCLS\Undulator\L143-112000-17\DATASET0001\Rough Tuning\Z Scans\Run 146\bscanz_dat.001
Temperature (C)
L143-112000-17:--> start 20.21 --> mid 20.18 --> end 20.06 --> bot 20.17 --> gra 20.17 --> air 20.04 
Axes position (m) X:0.027907 Y:0.000088
Initial Hall Probe Zero Measurements:
Vx_0i = -0.000400 V
Vy_0i = -0.000349 V
Final Hall Probe Zero Measurements:
Vx_0f = -0.000368 V
Vy_0f = -0.000370 V
Offset Correction Voff = offset+index*increment
index = 0,1...19850
Vx:offset = -0.000400 V, increment = 1.629E-09 V
Vy:offset = -0.000349 V, increment = -1.034E-09 V
Hall Probe Calibration File:
Y: C:\CVI\current_calib\SENIS_24-06\SENIS_24-06_y.ini
X: c:\cvi\current_calib\HALLx_calib_poly_fit_file.dat
Reference Magnet Measurements:
Bx_ref = 0.0000xx (T)
By_ref = 0.0000xx (T)
Magnetic Field Measurements vs Z

Analysis Parameters:
Input file directory = S:\MagData\LCLS\Undulator\L143-112000-17\DATASET0001\Rough Tuning\Z Scans\Run 146\
Input file name = bscanz_dat.001
Input file type = ZVXYBXYH
Correction required = 0
Correction file = S:\Magdata\Earth Field Correction\earth_field_corr.dat
Correction file type = CORR_ZBXY
Electron gamma = 26690.650000
Electron energy = 13.638897 Gev
Nominal undulator period = 0.030000 m
Nominal K value = 3.499557
Nominal radiation wavelength = 1.499903e-010 m
Number of undulator poles = 226
Number of end poles (used in analysis) = 4
Cell length = 3.656000 m

Peak And Zero Field Summary:
Number of Poles = 226
Number of End Poles (excluded in statistics) = 4
Peak Fields: Ave = 1.276746 T, RMS = 0.005337 T
Peak Field Spacing: Ave = 0.015001 m, RMS = 0.000023 m
Zero Field Spacing: Ave = 0.015001 m, RMS = 0.000033 m
Average Undulator Period = 0.030002 m

Hall Probe Overall Field Integrals:
Int Bx dz =   3.7419e-005 Tm
Int Int Bx dz dz =   7.2929e-005 Tm^2
Int By dz =   2.8003e-005 Tm
Int Int By dz dz =  -1.0849e-004 Tm^2

Trajectory Exit Information:
X at exit =  -2.3847e-006 m
X slope at exit =   6.1552e-007
Y at exit =  -1.6030e-006 m
Y slope at exit =  -8.2250e-007

Calculated Undulator Parameters:
Keff = 3.499865
Beff = 1.249358 T
Lambda_rad = 1.500208e-010 m

Phase Fluctuations:
Slippage vs Z At Peaks Fitted Slope =   5.00044e-009
Undulator Period = 0.030002 m
Radiation Wavelength =   1.50021e-010 m
Phase Error (From Slippage vs Z Fit Residuals): Ave = 0.000000 deg, RMS = 3.369418 deg

Phase Relative To Nominal Radiation Wave:
Nominal Radiation Wavelength =   1.49990e-010 m
Cell Length = 3.656000 m
Number Of End Poles = 4
Phase Change Into Undulator (To Pole After End Poles) =     824.119 deg, mod_360 =     104.119 deg
Phase Change Through Undulator (Poles Between End Poles) =   39073.605 deg, mod_360 =     193.605 deg
Phase Change Out Of Undulator (To Pole Before End Poles) =     795.371 deg, mod_360 =      75.371 deg
Phase Change Of Cell =   40693.096 deg
Nominal Phase Change Of Cell =   40680.000 deg
Phase Error Of Cell =      13.096 deg