SLAC Magnetic Measurements Date: 11-17-2009 Time: 09:07:52 Planar Insertion Device Analysis -------------------------------- Data File Header: SLAC Magnetic Measurements Date: 11-17-2009 Time: 09:02:05 Project: LCLS Device Type: Undulator Data Type: Serial Number: L143-112000-04 DATASET Number: 2 Tuning Stage: Fine Tuning Run Number: 10 Measurement Type: Z Scans Operator: YL Comment: find first pole position; large gap Session first run: 10 Measurement Device x: Measurement Device y: SENTRON_441 Data File: s:\magdata\LCLS\Undulator\L143-112000-04\DATASET0002\Fine Tuning\Z Scans\Run 010\bscanz.dat Temperature (C) L143-112000-04:--> T5 20.00 --> T2 19.99 --> T1 20.02 --> T3 19.99 --> Air 20.01 --> T4 20.00 Axes position (m) X:0.026532 Y:0.000213 Initial Hall Probe Zero Measurements: Vx_0i = 0.001436 V Vy_0i = -0.001964 V Final Hall Probe Zero Measurements: Vx_0f = 0.001537 V Vy_0f = -0.001822 V Offset Correction Voff = offset+index*increment index = 0,1...21250 Vx:offset = 0.001436 V, increment = 4.741E-09 V Vy:offset = -0.001964 V, increment = 6.690E-09 V Hall Probe Calibration File: Y: c:\CVI\current_calib\Sentron_441\Sentron_441_y.ini X: c:\cvi\current_calib\HALLx_calib_poly_fit_file.dat Reference Magnet Measurements: Bx_ref = 0.0000xx (T) By_ref = 0.0000xx (T) Magnetic Field Measurements vs Z Analysis Parameters: Input file directory = s:\magdata\LCLS\Undulator\L143-112000-04\DATASET0002\Fine Tuning\Z Scans\Run 010\ Input file name = x+00000_y+000_bscanz.dat Input file type = ZVXYBXYH Correction required = 1 Correction, Bx multiplicative factor = 1.000000 Correction, By multiplicative factor = 1.000000 Correction, Bx additive constant = -0.000007 T Correction, By additive constant = -0.000012 T Correction Bxy additive file = none Correction Bxy additive file type = CORR_ZBXY Electron gamma = 26625.060000 Electron energy = 13.605381 Gev Nominal undulator period = 0.030000 m Nominal K value = 3.485345 Nominal radiation wavelength = 1.496800e-010 m Number of undulator poles = 226 Number of end poles (used in analysis) = 4 Cell length = 3.656000 m Nominal phase change through cell = 709.999940 rad Nominal phase of pole after end poles = 14.137167 rad Save as final result = 0 Result file directory = c:\magdata\slac_und_2\ Result file name = zscan_xn0060yp0002.dat Figures close on exit = 0 Figures visible = 1 !!!!!!!!!!!!!!!!!!!!!!!!!!!!!! The Measured Fields Were Corrected: Bx_corr = CORR_BX_MULT * Bx + CORR_BX_ADD + CORR_BX_ADD_FILE By_corr = CORR_BY_MULT * By + CORR_BY_ADD + CORR_BY_ADD_FILE Parameters: CORR_BX_MULT = 1.000000 CORR_BY_MULT = 1.000000 CORR_BX_ADD = -0.000007 T CORR_BY_ADD = -0.000012 T CORR_BX_ADD_FILE = none CORR_BY_ADD_FILE = none !!!!!!!!!!!!!!!!!!!!!!!!!!!!!! Peak And Zero Field Summary: Number of Poles = 226 Number of End Poles (excluded in statistics) = 4 Peak Fields: Ave = 0.826882 T, RMS = 0.002767 T Peak Field Spacing: Ave = 0.015001 m, RMS = 0.000029 m Zero Field Spacing: Ave = 0.015001 m, RMS = 0.000033 m Average Undulator Period = 0.030001 m Hall Probe Overall Field Integrals: Int Bx dz = 1.6205e-005 Tm Int Int Bx dz dz = 5.5773e-005 Tm^2 Int By dz = -4.9472e-005 Tm Int Int By dz dz = 8.1678e-005 Tm^2 Trajectory Exit Information: X at exit = 1.7998e-006 m X slope at exit = -1.0901e-006 Y at exit = -1.2289e-006 m Y slope at exit = -3.5707e-007 Calculated Undulator Parameters: Keff = 2.297410 Beff = 0.820140 T Lambda_rad = 7.700275e-011 m Calculated K Value At 20 C: (Assumes Keff changes by -.05% per degree C) Measured Average Undulator Temperature = 20.000000 C Measured Keff = 2.297410 Calculated Keff at 20 C = 2.297410 Relative Change (Keff_20C - Keff) / Keff = 0.000000 Phase Fluctuations: Slippage vs Z At Peaks Fitted Slope = 2.56671e-009 Undulator Period = 0.030001 m Radiation Wavelength = 7.70028e-011 m Phase Error (From Slippage vs Z Fit Residuals): Ave = 0.000000 deg, Phase error RMS = 12.130575 deg Phase Relative To Nominal Radiation Wave: Nominal Radiation Wavelength = 1.49680e-010 m Cell Length = 3.656000 m Number Of End Poles = 4 Entrance: Phase Error Of First 10 Poles After The End Poles Ave = -632.350556 deg, RMS = 265.615547 deg Center: Phase Error Of All Poles In Undulator Center: Ave = -9709.223966 deg, RMS = 5512.559475 deg Exit: Phase Error To Cell Boundary Of Last 10 Poles Before The End Poles Ave = -624.647610 deg, RMS = 263.147652 deg Cell: Phase Change Of Cell = 21250.194 deg Nominal Phase Change Of Cell = 40680.000 deg Phase Error Of Cell = -19429.806 deg Calculated Phase Change In The Cell At 20 C: (Relative to the nominal radiation wave) (Assumes Keff changes by -.05% per degree C) Measured Average Undulator Temperature = 20.000000 C Measured Phase Change Of Cell = 21250.194 deg Calculated Phase Change Of Cell At 20 C = 21250.194 deg Phase Change Due To Temperature (Phase_20C - Phase) = 0.000 deg