SLAC Magnetic Measurements Date: 10-15-2008 Time: 11:20:44 Planar Insertion Device Analysis -------------------------------- Data File Header: SLAC Magnetic Measurements Date: 10-15-2008 Time: 10:56:36 Project: LCLS Device Type: Undulator Data Type: Serial Number: L143-112000-03 DATASET Number: 2 Tuning Stage: Fine Tuning Run Number: 16 Measurement Type: Z Scans Operator: YL Comment: final scans Session first run: 15 Measurement Device x: Measurement Device y: SENTRON_441 Data File: s:\magdata\LCLS\Undulator\L143-112000-03\DATASET0002\Fine Tuning\Z Scans\Run 016\bscanz.dat Temperature (C) L143-112000-03:--> top_end 19.89 --> top_mid 19.91 --> top_start 19.93 --> bottom 19.91 --> Air 19.96 --> Granite 19.90 Axes position (m) X:0.027715 Y:0.000305 Initial Hall Probe Zero Measurements: Vx_0i = 0.001775 V Vy_0i = -0.000618 V Final Hall Probe Zero Measurements: Vx_0f = 0.001773 V Vy_0f = -0.000629 V Offset Correction Voff = offset+index*increment index = 0,1...21250 Vx:offset = 0.001775 V, increment = -7.843E-11 V Vy:offset = -0.000618 V, increment = -5.541E-10 V Hall Probe Calibration File: Y: c:\CVI\current_calib\Sentron_441\Sentron_441_y.ini X: c:\cvi\current_calib\HALLx_calib_poly_fit_file.dat Reference Magnet Measurements: Bx_ref = 0.0000xx (T) By_ref = 0.0000xx (T) Magnetic Field Measurements vs Z Analysis Parameters: Input file directory = s:\MagData\LCLS\Undulator\L143-112000-03\DATASET0002\Fine Tuning\Z Scans\Run 016\ Input file name = x+00000_y+000_bscanz.dat Input file type = ZVXYBXYH Correction required = 1 Correction, Bx multiplicative factor = 1.000000 Correction, By multiplicative factor = 1.000000 Correction, Bx additive constant = -0.000010 T Correction, By additive constant = -0.000006 T Correction Bxy additive file = none Correction Bxy additive file type = CORR_ZBXY Electron gamma = 26641.850000 Electron energy = 13.613960 Gev Nominal undulator period = 0.030000 m Nominal K value = 3.488984 Nominal radiation wavelength = 1.497596e-010 m Number of undulator poles = 226 Number of end poles (used in analysis) = 4 Cell length = 3.656000 m Nominal phase change through cell = 709.999940 rad Nominal phase of pole after end poles = 14.137167 rad Save as final result = 0 Result file directory = c:\magdata\slac_und_2\ Result file name = zscan_xn0060yp0002.dat Figures close on exit = 0 Figures visible = 1 !!!!!!!!!!!!!!!!!!!!!!!!!!!!!! The Measured Fields Were Corrected: Bx_corr = CORR_BX_MULT * Bx + CORR_BX_ADD + CORR_BX_ADD_FILE By_corr = CORR_BY_MULT * By + CORR_BY_ADD + CORR_BY_ADD_FILE Parameters: CORR_BX_MULT = 1.000000 CORR_BY_MULT = 1.000000 CORR_BX_ADD = -0.000010 T CORR_BY_ADD = -0.000006 T CORR_BX_ADD_FILE = none CORR_BY_ADD_FILE = none !!!!!!!!!!!!!!!!!!!!!!!!!!!!!! Peak And Zero Field Summary: Number of Poles = 226 Number of End Poles (excluded in statistics) = 4 Peak Fields: Ave = 1.273334 T, RMS = 0.006345 T Peak Field Spacing: Ave = 0.015000 m, RMS = 0.000024 m Zero Field Spacing: Ave = 0.015000 m, RMS = 0.000036 m Average Undulator Period = 0.030001 m Hall Probe Overall Field Integrals: Int Bx dz = -2.6191e-005 Tm Int Int Bx dz dz = 9.1200e-006 Tm^2 Int By dz = 1.5949e-005 Tm Int Int By dz dz = 1.3831e-005 Tm^2 Trajectory Exit Information: X at exit = 3.0456e-007 m X slope at exit = 3.5121e-007 Y at exit = -2.0083e-007 m Y slope at exit = 5.7676e-007 Calculated Undulator Parameters: Keff = 3.488982 Beff = 1.245514 T Lambda_rad = 1.497623e-010 m Calculated K Value At 20 C: (Assumes Keff changes by -.05% per degree C) Measured Average Undulator Temperature = 19.910000 C Measured Keff = 3.488982 Calculated Keff at 20 C = 3.488825 Relative Change (Keff_20C - Keff) / Keff = -0.000045 Phase Fluctuations: Slippage vs Z At Peaks Fitted Slope = 4.99198e-009 Undulator Period = 0.030001 m Radiation Wavelength = 1.49762e-010 m Phase Error (From Slippage vs Z Fit Residuals): Ave = 0.000000 deg, Phase error RMS = 3.115703 deg Phase Relative To Nominal Radiation Wave: Nominal Radiation Wavelength = 1.49760e-010 m Cell Length = 3.656000 m Number Of End Poles = 4 Entrance: Phase Error Of First 10 Poles After The End Poles Ave = -1.391393 deg, RMS = 1.962125 deg Center: Phase Error Of All Poles In Undulator Center: Ave = -1.659561 deg, RMS = 3.147476 deg Exit: Phase Error To Cell Boundary Of Last 10 Poles Before The End Poles Ave = -0.113241 deg, RMS = 4.016982 deg Cell: Phase Change Of Cell = 40678.965 deg Nominal Phase Change Of Cell = 40680.000 deg Phase Error Of Cell = -1.035 deg Calculated Phase Change In The Cell At 20 C: (Relative to the nominal radiation wave) (Assumes Keff changes by -.05% per degree C) Measured Average Undulator Temperature = 19.910000 C Measured Phase Change Of Cell = 40678.965 deg Calculated Phase Change Of Cell At 20 C = 40675.821 deg Phase Change Due To Temperature (Phase_20C - Phase) = -3.144 deg