SLAC Magnetic Measurements Date: 06-14-2017 Time: 17:21:47 Project: LCLS_II Magnet Type: Solenoid Magnet Name: SOL1 Serial Number: Run Number: 15 Measurement Device: LS475 Operator: sda Comment: Bz field vs current from -10 to 10 A in 0.5 A steps at SOL1 mechanical center. Standardization for Main: 2 cycles from -10.0 A to 10.0 A to -10.0 A Test Currents for Main (A): -10.00 -9.50 -9.00 -8.50 -8.00 -7.50 -7.00 -6.50 -6.00 -5.50 -5.00 -4.50 -4.00 -3.50 -3.00 -2.50 -2.00 -1.50 -1.00 -0.50 0.00 0.50 1.00 1.50 2.00 2.50 3.00 3.50 4.00 4.50 5.00 5.50 6.00 6.50 7.00 7.50 8.00 8.50 9.00 9.50 10.00 17:21:47 Stage Position Measurement, pos = 0.0000000 m 17:21:51 Move Stage, position = 0.000000 17:21:51 MoveZ Devices: Move Device = MANUAL, Read Device = NONE 17:22:27 Move Stage, position = 0.440000 17:23:05 Hall Probe Zeroed 17:24:06 Hall Probe Type Being Used, type = LS475 17:24:08 Move Stage, position = 0.000000 17:24:19 Magnet Current for Main, Imag = -0.0003 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = 0.000000 A 17:24:19 Magnet Current for Main, Imag = -0.0003 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = 0.000000 A 17:24:20 Magnet Current for Main, Imag = -0.0003 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = 0.000000 A 17:24:20 Magnet Current for Main, Imag = -0.0003 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = 0.000000 A 17:24:20 Magnet Avgerage Current Measurement for Main : Imag = -0.0003 +- 0.0000 A 17:24:32 Initial Zero Current Measurement for Main, Imag_zero = -0.000285597 A 17:24:32 Hall Probe Range Set, range = 3 17:24:32 Magnet system Main current ramping to 10.0000 A... 17:24:47 Magnet system Main current ramping to -10.0000 A... 17:25:07 Magnet system Main current ramping to 10.0000 A... 17:25:28 Magnet system Main current ramping to -10.0000 A... 17:25:48 Magnet system Main current ramping to -10.0000 A... 17:26:02 Move Stage, position = 0.000000 17:26:04 Stage Position Measurement, pos = 0.0000000 m 17:26:05 Magnet Current for Main, Imag = -10.0090 A Vtrans = -2.001860 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000286 A 17:26:06 Magnet Current for Main, Imag = -10.0090 A Vtrans = -2.001856 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000286 A 17:26:06 Magnet Current for Main, Imag = -10.0090 A Vtrans = -2.001855 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000286 A 17:26:07 Magnet Current for Main, Imag = -10.0090 A Vtrans = -2.001856 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000286 A 17:26:07 Magnet Avgerage Current Measurement for Main : Imag = -10.0090 +- 0.0000 A 17:26:07 Hall Probe Field Measurement, B = -0.094837 T 17:26:08 Hall Probe Field Measurement, B = -0.094837 T 17:26:08 Hall Probe Field Measurement, B = -0.094836 T 17:26:09 Hall Probe Field Measurement, B = -0.094837 T 17:26:09 Hall Probe Field Measurement, B = -0.094837 T 17:26:09 Hall Probe Field Measurement, B = -0.094838 T 17:26:10 Hall Probe Field Measurement: B = -0.094837 +/- 0.000001 T 17:26:10 Move Stage, position = 0.000000 17:26:10 Magnet system Main current ramping to -9.5000 A... 17:26:20 Move Stage, position = 0.000000 17:26:22 Stage Position Measurement, pos = 0.0000000 m 17:26:23 Magnet Current for Main, Imag = -9.5057 A Vtrans = -1.901193 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000286 A 17:26:24 Magnet Current for Main, Imag = -9.5057 A Vtrans = -1.901199 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000286 A 17:26:25 Magnet Current for Main, Imag = -9.5057 A Vtrans = -1.901196 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000286 A 17:26:25 Magnet Current for Main, Imag = -9.5057 A Vtrans = -1.901197 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000286 A 17:26:25 Magnet Avgerage Current Measurement for Main : Imag = -9.5057 +- 0.0000 A 17:26:25 Hall Probe Field Measurement, B = -0.090073 T 17:26:26 Hall Probe Field Measurement, B = -0.090073 T 17:26:26 Hall Probe Field Measurement, B = -0.090073 T 17:26:26 Hall Probe Field Measurement, B = -0.090073 T 17:26:27 Hall Probe Field Measurement, B = -0.090073 T 17:26:27 Hall Probe Field Measurement, B = -0.090074 T 17:26:28 Hall Probe Field Measurement: B = -0.090073 +/- 0.000000 T 17:26:28 Move Stage, position = 0.000000 17:26:28 Magnet system Main current ramping to -9.0000 A... 17:26:38 Move Stage, position = 0.000000 17:26:40 Stage Position Measurement, pos = 0.0000000 m 17:26:41 Magnet Current for Main, Imag = -9.0097 A Vtrans = -1.801997 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000286 A 17:26:42 Magnet Current for Main, Imag = -9.0097 A Vtrans = -1.801996 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000286 A 17:26:42 Magnet Current for Main, Imag = -9.0097 A Vtrans = -1.801996 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000286 A 17:26:43 Magnet Current for Main, Imag = -9.0097 A Vtrans = -1.801996 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000286 A 17:26:43 Magnet Avgerage Current Measurement for Main : Imag = -9.0097 +- 0.0000 A 17:26:43 Hall Probe Field Measurement, B = -0.085382 T 17:26:44 Hall Probe Field Measurement, B = -0.085381 T 17:26:44 Hall Probe Field Measurement, B = -0.085381 T 17:26:45 Hall Probe Field Measurement, B = -0.085381 T 17:26:45 Hall Probe Field Measurement, B = -0.085381 T 17:26:45 Hall Probe Field Measurement, B = -0.085381 T 17:26:46 Hall Probe Field Measurement: B = -0.085381 +/- 0.000000 T 17:26:46 Move Stage, position = 0.000000 17:26:46 Magnet system Main current ramping to -8.5000 A... 17:26:56 Move Stage, position = 0.000000 17:26:58 Stage Position Measurement, pos = 0.0000000 m 17:26:59 Magnet Current for Main, Imag = -8.5077 A Vtrans = -1.701596 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000286 A 17:27:00 Magnet Current for Main, Imag = -8.5077 A Vtrans = -1.701591 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000286 A 17:27:01 Magnet Current for Main, Imag = -8.5077 A Vtrans = -1.701593 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000286 A 17:27:01 Magnet Current for Main, Imag = -8.5077 A Vtrans = -1.701593 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000286 A 17:27:01 Magnet Avgerage Current Measurement for Main : Imag = -8.5077 +- 0.0000 A 17:27:01 Hall Probe Field Measurement, B = -0.080623 T 17:27:02 Hall Probe Field Measurement, B = -0.080623 T 17:27:02 Hall Probe Field Measurement, B = -0.080624 T 17:27:03 Hall Probe Field Measurement, B = -0.080624 T 17:27:03 Hall Probe Field Measurement, B = -0.080624 T 17:27:03 Hall Probe Field Measurement, B = -0.080624 T 17:27:04 Hall Probe Field Measurement: B = -0.080624 +/- 0.000000 T 17:27:04 Move Stage, position = 0.000000 17:27:04 Magnet system Main current ramping to -8.0000 A... 17:27:14 Move Stage, position = 0.000000 17:27:16 Stage Position Measurement, pos = 0.0000000 m 17:27:17 Magnet Current for Main, Imag = -8.0085 A Vtrans = -1.601756 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000286 A 17:27:18 Magnet Current for Main, Imag = -8.0085 A Vtrans = -1.601756 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000286 A 17:27:19 Magnet Current for Main, Imag = -8.0085 A Vtrans = -1.601756 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000286 A 17:27:19 Magnet Current for Main, Imag = -8.0085 A Vtrans = -1.601753 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000286 A 17:27:19 Magnet Avgerage Current Measurement for Main : Imag = -8.0085 +- 0.0000 A 17:27:19 Hall Probe Field Measurement, B = -0.075900 T 17:27:20 Hall Probe Field Measurement, B = -0.075900 T 17:27:20 Hall Probe Field Measurement, B = -0.075900 T 17:27:21 Hall Probe Field Measurement, B = -0.075900 T 17:27:21 Hall Probe Field Measurement, B = -0.075900 T 17:27:21 Hall Probe Field Measurement, B = -0.075900 T 17:27:22 Hall Probe Field Measurement: B = -0.075900 +/- 0.000000 T 17:27:22 Move Stage, position = 0.000000 17:27:22 Magnet system Main current ramping to -7.5000 A... 17:27:32 Move Stage, position = 0.000000 17:27:34 Stage Position Measurement, pos = 0.0000000 m 17:27:35 Magnet Current for Main, Imag = -7.5082 A Vtrans = -1.501691 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000286 A 17:27:36 Magnet Current for Main, Imag = -7.5081 A Vtrans = -1.501684 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000286 A 17:27:37 Magnet Current for Main, Imag = -7.5082 A Vtrans = -1.501692 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000286 A 17:27:37 Magnet Current for Main, Imag = -7.5081 A Vtrans = -1.501687 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000286 A 17:27:37 Magnet Avgerage Current Measurement for Main : Imag = -7.5082 +- 0.0000 A 17:27:37 Hall Probe Field Measurement, B = -0.071159 T 17:27:38 Hall Probe Field Measurement, B = -0.071159 T 17:27:38 Hall Probe Field Measurement, B = -0.071159 T 17:27:39 Hall Probe Field Measurement, B = -0.071159 T 17:27:39 Hall Probe Field Measurement, B = -0.071159 T 17:27:39 Hall Probe Field Measurement, B = -0.071159 T 17:27:40 Hall Probe Field Measurement: B = -0.071159 +/- 0.000000 T 17:27:40 Move Stage, position = 0.000000 17:27:40 Magnet system Main current ramping to -7.0000 A... 17:27:50 Move Stage, position = 0.000000 17:27:52 Stage Position Measurement, pos = 0.0000000 m 17:27:53 Magnet Current for Main, Imag = -7.0047 A Vtrans = -1.401003 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000286 A 17:27:54 Magnet Current for Main, Imag = -7.0047 A Vtrans = -1.400995 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000286 A 17:27:55 Magnet Current for Main, Imag = -7.0047 A Vtrans = -1.401005 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000286 A 17:27:55 Magnet Current for Main, Imag = -7.0047 A Vtrans = -1.401002 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000286 A 17:27:55 Magnet Avgerage Current Measurement for Main : Imag = -7.0047 +- 0.0000 A 17:27:55 Hall Probe Field Measurement, B = -0.066389 T 17:27:56 Hall Probe Field Measurement, B = -0.066389 T 17:27:56 Hall Probe Field Measurement, B = -0.066389 T 17:27:57 Hall Probe Field Measurement, B = -0.066389 T 17:27:57 Hall Probe Field Measurement, B = -0.066388 T 17:27:57 Hall Probe Field Measurement, B = -0.066388 T 17:27:58 Hall Probe Field Measurement: B = -0.066389 +/- 0.000000 T 17:27:58 Move Stage, position = 0.000000 17:27:58 Magnet system Main current ramping to -6.5000 A... 17:28:08 Move Stage, position = 0.000000 17:28:10 Stage Position Measurement, pos = 0.0000000 m 17:28:11 Magnet Current for Main, Imag = -6.5059 A Vtrans = -1.301243 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000286 A 17:28:12 Magnet Current for Main, Imag = -6.5059 A Vtrans = -1.301246 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000286 A 17:28:13 Magnet Current for Main, Imag = -6.5059 A Vtrans = -1.301240 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000286 A 17:28:13 Magnet Current for Main, Imag = -6.5059 A Vtrans = -1.301242 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000286 A 17:28:13 Magnet Avgerage Current Measurement for Main : Imag = -6.5059 +- 0.0000 A 17:28:13 Hall Probe Field Measurement, B = -0.061664 T 17:28:14 Hall Probe Field Measurement, B = -0.061664 T 17:28:14 Hall Probe Field Measurement, B = -0.061664 T 17:28:15 Hall Probe Field Measurement, B = -0.061664 T 17:28:15 Hall Probe Field Measurement, B = -0.061663 T 17:28:15 Hall Probe Field Measurement, B = -0.061663 T 17:28:16 Hall Probe Field Measurement: B = -0.061664 +/- 0.000000 T 17:28:16 Move Stage, position = 0.000000 17:28:16 Magnet system Main current ramping to -6.0000 A... 17:28:26 Move Stage, position = 0.000000 17:28:28 Stage Position Measurement, pos = 0.0000000 m 17:28:29 Magnet Current for Main, Imag = -6.0093 A Vtrans = -1.201907 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000286 A 17:28:30 Magnet Current for Main, Imag = -6.0092 A Vtrans = -1.201906 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000286 A 17:28:31 Magnet Current for Main, Imag = -6.0093 A Vtrans = -1.201911 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000286 A 17:28:31 Magnet Current for Main, Imag = -6.0093 A Vtrans = -1.201911 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000286 A 17:28:31 Magnet Avgerage Current Measurement for Main : Imag = -6.0093 +- 0.0000 A 17:28:31 Hall Probe Field Measurement, B = -0.056959 T 17:28:32 Hall Probe Field Measurement, B = -0.056960 T 17:28:32 Hall Probe Field Measurement, B = -0.056960 T 17:28:33 Hall Probe Field Measurement, B = -0.056961 T 17:28:33 Hall Probe Field Measurement, B = -0.056961 T 17:28:34 Hall Probe Field Measurement, B = -0.056961 T 17:28:34 Hall Probe Field Measurement: B = -0.056960 +/- 0.000001 T 17:28:34 Move Stage, position = 0.000000 17:28:34 Magnet system Main current ramping to -5.5000 A... 17:28:45 Move Stage, position = 0.000000 17:28:47 Stage Position Measurement, pos = 0.0000000 m 17:28:47 Magnet Current for Main, Imag = -5.5074 A Vtrans = -1.101534 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000286 A 17:28:48 Magnet Current for Main, Imag = -5.5074 A Vtrans = -1.101533 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000286 A 17:28:49 Magnet Current for Main, Imag = -5.5074 A Vtrans = -1.101533 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000286 A 17:28:49 Magnet Current for Main, Imag = -5.5074 A Vtrans = -1.101533 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000286 A 17:28:49 Magnet Avgerage Current Measurement for Main : Imag = -5.5074 +- 0.0000 A 17:28:49 Hall Probe Field Measurement, B = -0.052212 T 17:28:50 Hall Probe Field Measurement, B = -0.052212 T 17:28:50 Hall Probe Field Measurement, B = -0.052213 T 17:28:51 Hall Probe Field Measurement, B = -0.052213 T 17:28:51 Hall Probe Field Measurement, B = -0.052213 T 17:28:52 Hall Probe Field Measurement, B = -0.052213 T 17:28:52 Hall Probe Field Measurement: B = -0.052213 +/- 0.000000 T 17:28:52 Move Stage, position = 0.000000 17:28:52 Magnet system Main current ramping to -5.0000 A... 17:29:03 Move Stage, position = 0.000000 17:29:05 Stage Position Measurement, pos = 0.0000000 m 17:29:05 Magnet Current for Main, Imag = -5.0088 A Vtrans = -1.001824 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000286 A 17:29:06 Magnet Current for Main, Imag = -5.0088 A Vtrans = -1.001823 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000286 A 17:29:07 Magnet Current for Main, Imag = -5.0088 A Vtrans = -1.001822 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000286 A 17:29:07 Magnet Current for Main, Imag = -5.0088 A Vtrans = -1.001819 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000286 A 17:29:07 Magnet Avgerage Current Measurement for Main : Imag = -5.0088 +- 0.0000 A 17:29:07 Hall Probe Field Measurement, B = -0.047490 T 17:29:08 Hall Probe Field Measurement, B = -0.047489 T 17:29:08 Hall Probe Field Measurement, B = -0.047489 T 17:29:09 Hall Probe Field Measurement, B = -0.047489 T 17:29:09 Hall Probe Field Measurement, B = -0.047489 T 17:29:10 Hall Probe Field Measurement, B = -0.047489 T 17:29:10 Hall Probe Field Measurement: B = -0.047489 +/- 0.000000 T 17:29:10 Move Stage, position = 0.000000 17:29:10 Magnet system Main current ramping to -4.5000 A... 17:29:21 Move Stage, position = 0.000000 17:29:23 Stage Position Measurement, pos = 0.0000000 m 17:29:23 Magnet Current for Main, Imag = -4.5063 A Vtrans = -0.901314 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000286 A 17:29:24 Magnet Current for Main, Imag = -4.5063 A Vtrans = -0.901313 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000286 A 17:29:25 Magnet Current for Main, Imag = -4.5063 A Vtrans = -0.901315 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000286 A 17:29:25 Magnet Current for Main, Imag = -4.5063 A Vtrans = -0.901312 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000286 A 17:29:25 Magnet Avgerage Current Measurement for Main : Imag = -4.5063 +- 0.0000 A 17:29:25 Hall Probe Field Measurement, B = -0.042727 T 17:29:26 Hall Probe Field Measurement, B = -0.042728 T 17:29:26 Hall Probe Field Measurement, B = -0.042728 T 17:29:27 Hall Probe Field Measurement, B = -0.042727 T 17:29:27 Hall Probe Field Measurement, B = -0.042727 T 17:29:28 Hall Probe Field Measurement, B = -0.042728 T 17:29:28 Hall Probe Field Measurement: B = -0.042728 +/- 0.000001 T 17:29:28 Move Stage, position = 0.000000 17:29:28 Magnet system Main current ramping to -4.0000 A... 17:29:39 Move Stage, position = 0.000000 17:29:41 Stage Position Measurement, pos = 0.0000000 m 17:29:41 Magnet Current for Main, Imag = -4.0050 A Vtrans = -0.801054 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000286 A 17:29:42 Magnet Current for Main, Imag = -4.0050 A Vtrans = -0.801051 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000286 A 17:29:43 Magnet Current for Main, Imag = -4.0050 A Vtrans = -0.801052 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000286 A 17:29:43 Magnet Current for Main, Imag = -4.0050 A Vtrans = -0.801050 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000286 A 17:29:43 Magnet Avgerage Current Measurement for Main : Imag = -4.0050 +- 0.0000 A 17:29:43 Hall Probe Field Measurement, B = -0.037984 T 17:29:44 Hall Probe Field Measurement, B = -0.037984 T 17:29:44 Hall Probe Field Measurement, B = -0.037984 T 17:29:45 Hall Probe Field Measurement, B = -0.037984 T 17:29:45 Hall Probe Field Measurement, B = -0.037984 T 17:29:46 Hall Probe Field Measurement, B = -0.037984 T 17:29:46 Hall Probe Field Measurement: B = -0.037984 +/- 0.000000 T 17:29:46 Move Stage, position = 0.000000 17:29:46 Magnet system Main current ramping to -3.5000 A... 17:29:57 Move Stage, position = 0.000000 17:29:59 Stage Position Measurement, pos = 0.0000000 m 17:29:59 Magnet Current for Main, Imag = -3.5056 A Vtrans = -0.701180 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000286 A 17:30:00 Magnet Current for Main, Imag = -3.5056 A Vtrans = -0.701179 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000286 A 17:30:01 Magnet Current for Main, Imag = -3.5056 A Vtrans = -0.701178 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000286 A 17:30:01 Magnet Current for Main, Imag = -3.5056 A Vtrans = -0.701181 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000286 A 17:30:01 Magnet Avgerage Current Measurement for Main : Imag = -3.5056 +- 0.0000 A 17:30:01 Hall Probe Field Measurement, B = -0.033257 T 17:30:02 Hall Probe Field Measurement, B = -0.033257 T 17:30:02 Hall Probe Field Measurement, B = -0.033257 T 17:30:03 Hall Probe Field Measurement, B = -0.033257 T 17:30:03 Hall Probe Field Measurement, B = -0.033257 T 17:30:04 Hall Probe Field Measurement, B = -0.033256 T 17:30:04 Hall Probe Field Measurement: B = -0.033257 +/- 0.000000 T 17:30:04 Move Stage, position = 0.000000 17:30:04 Magnet system Main current ramping to -3.0000 A... 17:30:15 Move Stage, position = 0.000000 17:30:17 Stage Position Measurement, pos = 0.0000000 m 17:30:17 Magnet Current for Main, Imag = -3.0081 A Vtrans = -0.601674 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000286 A 17:30:18 Magnet Current for Main, Imag = -3.0081 A Vtrans = -0.601673 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000286 A 17:30:19 Magnet Current for Main, Imag = -3.0081 A Vtrans = -0.601671 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000286 A 17:30:20 Magnet Current for Main, Imag = -3.0081 A Vtrans = -0.601674 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000286 A 17:30:20 Magnet Avgerage Current Measurement for Main : Imag = -3.0081 +- 0.0000 A 17:30:20 Hall Probe Field Measurement, B = -0.028555 T 17:30:20 Hall Probe Field Measurement, B = -0.028555 T 17:30:20 Hall Probe Field Measurement, B = -0.028554 T 17:30:21 Hall Probe Field Measurement, B = -0.028554 T 17:30:21 Hall Probe Field Measurement, B = -0.028553 T 17:30:22 Hall Probe Field Measurement, B = -0.028552 T 17:30:22 Hall Probe Field Measurement: B = -0.028554 +/- 0.000001 T 17:30:22 Move Stage, position = 0.000000 17:30:22 Magnet system Main current ramping to -2.5000 A... 17:30:33 Move Stage, position = 0.000000 17:30:35 Stage Position Measurement, pos = 0.0000000 m 17:30:35 Magnet Current for Main, Imag = -2.5059 A Vtrans = -0.501231 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000286 A 17:30:36 Magnet Current for Main, Imag = -2.5059 A Vtrans = -0.501230 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000286 A 17:30:37 Magnet Current for Main, Imag = -2.5059 A Vtrans = -0.501232 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000286 A 17:30:38 Magnet Current for Main, Imag = -2.5059 A Vtrans = -0.501230 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000286 A 17:30:38 Magnet Avgerage Current Measurement for Main : Imag = -2.5059 +- 0.0000 A 17:30:38 Hall Probe Field Measurement, B = -0.023797 T 17:30:38 Hall Probe Field Measurement, B = -0.023797 T 17:30:38 Hall Probe Field Measurement, B = -0.023797 T 17:30:39 Hall Probe Field Measurement, B = -0.023797 T 17:30:39 Hall Probe Field Measurement, B = -0.023797 T 17:30:40 Hall Probe Field Measurement, B = -0.023796 T 17:30:40 Hall Probe Field Measurement: B = -0.023797 +/- 0.000000 T 17:30:40 Move Stage, position = 0.000000 17:30:40 Magnet system Main current ramping to -2.0000 A... 17:30:51 Move Stage, position = 0.000000 17:30:53 Stage Position Measurement, pos = 0.0000000 m 17:30:53 Magnet Current for Main, Imag = -2.0007 A Vtrans = -0.400204 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000286 A 17:30:54 Magnet Current for Main, Imag = -2.0007 A Vtrans = -0.400205 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000286 A 17:30:55 Magnet Current for Main, Imag = -2.0007 A Vtrans = -0.400206 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000286 A 17:30:56 Magnet Current for Main, Imag = -2.0007 A Vtrans = -0.400206 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000286 A 17:30:56 Magnet Avgerage Current Measurement for Main : Imag = -2.0007 +- 0.0000 A 17:30:56 Hall Probe Field Measurement, B = -0.019005 T 17:30:56 Hall Probe Field Measurement, B = -0.019005 T 17:30:56 Hall Probe Field Measurement, B = -0.019006 T 17:30:57 Hall Probe Field Measurement, B = -0.019005 T 17:30:57 Hall Probe Field Measurement, B = -0.019005 T 17:30:58 Hall Probe Field Measurement, B = -0.019004 T 17:30:58 Hall Probe Field Measurement: B = -0.019005 +/- 0.000001 T 17:30:58 Move Stage, position = 0.000000 17:30:58 Magnet system Main current ramping to -1.5000 A... 17:31:09 Move Stage, position = 0.000000 17:31:11 Stage Position Measurement, pos = 0.0000000 m 17:31:11 Magnet Current for Main, Imag = -1.5055 A Vtrans = -0.301160 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000286 A 17:31:12 Magnet Current for Main, Imag = -1.5055 A Vtrans = -0.301160 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000286 A 17:31:13 Magnet Current for Main, Imag = -1.5055 A Vtrans = -0.301160 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000286 A 17:31:14 Magnet Current for Main, Imag = -1.5055 A Vtrans = -0.301160 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000286 A 17:31:14 Magnet Avgerage Current Measurement for Main : Imag = -1.5055 +- 0.0000 A 17:31:14 Hall Probe Field Measurement, B = -0.014313 T 17:31:14 Hall Probe Field Measurement, B = -0.014313 T 17:31:14 Hall Probe Field Measurement, B = -0.014313 T 17:31:15 Hall Probe Field Measurement, B = -0.014313 T 17:31:15 Hall Probe Field Measurement, B = -0.014312 T 17:31:16 Hall Probe Field Measurement, B = -0.014312 T 17:31:16 Hall Probe Field Measurement: B = -0.014313 +/- 0.000000 T 17:31:16 Move Stage, position = 0.000000 17:31:16 Magnet system Main current ramping to -1.0000 A... 17:31:27 Move Stage, position = 0.000000 17:31:29 Stage Position Measurement, pos = 0.0000000 m 17:31:30 Magnet Current for Main, Imag = -1.0026 A Vtrans = -0.200580 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000286 A 17:31:30 Magnet Current for Main, Imag = -1.0026 A Vtrans = -0.200578 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000286 A 17:31:31 Magnet Current for Main, Imag = -1.0026 A Vtrans = -0.200575 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000286 A 17:31:32 Magnet Current for Main, Imag = -1.0026 A Vtrans = -0.200575 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000286 A 17:31:32 Magnet Avgerage Current Measurement for Main : Imag = -1.0026 +- 0.0000 A 17:31:32 Hall Probe Field Measurement, B = -0.009541 T 17:31:32 Hall Probe Field Measurement, B = -0.009541 T 17:31:33 Hall Probe Field Measurement, B = -0.009540 T 17:31:33 Hall Probe Field Measurement, B = -0.009540 T 17:31:33 Hall Probe Field Measurement, B = -0.009539 T 17:31:34 Hall Probe Field Measurement, B = -0.009539 T 17:31:34 Hall Probe Field Measurement: B = -0.009540 +/- 0.000001 T 17:31:34 Move Stage, position = 0.000000 17:31:34 Magnet system Main current ramping to -0.5000 A... 17:31:45 Move Stage, position = 0.000000 17:31:47 Stage Position Measurement, pos = 0.0000000 m 17:31:48 Magnet Current for Main, Imag = -0.5021 A Vtrans = -0.100474 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000286 A 17:31:48 Magnet Current for Main, Imag = -0.5021 A Vtrans = -0.100475 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000286 A 17:31:49 Magnet Current for Main, Imag = -0.5021 A Vtrans = -0.100475 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000286 A 17:31:50 Magnet Current for Main, Imag = -0.5021 A Vtrans = -0.100477 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000286 A 17:31:50 Magnet Avgerage Current Measurement for Main : Imag = -0.5021 +- 0.0000 A 17:31:50 Hall Probe Field Measurement, B = -0.004798 T 17:31:50 Hall Probe Field Measurement, B = -0.004798 T 17:31:51 Hall Probe Field Measurement, B = -0.004799 T 17:31:51 Hall Probe Field Measurement, B = -0.004798 T 17:31:51 Hall Probe Field Measurement, B = -0.004798 T 17:31:52 Hall Probe Field Measurement, B = -0.004797 T 17:31:52 Hall Probe Field Measurement: B = -0.004798 +/- 0.000001 T 17:31:52 Move Stage, position = 0.000000 17:31:52 Magnet system Main current ramping to 0.0000 A... 17:32:03 Move Stage, position = 0.000000 17:32:05 Stage Position Measurement, pos = 0.0000000 m 17:32:06 Magnet Current for Main, Imag = -0.0020 A Vtrans = -0.000462 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000286 A 17:32:06 Magnet Current for Main, Imag = -0.0020 A Vtrans = -0.000465 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000286 A 17:32:07 Magnet Current for Main, Imag = -0.0020 A Vtrans = -0.000463 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000286 A 17:32:08 Magnet Current for Main, Imag = -0.0020 A Vtrans = -0.000465 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000286 A 17:32:08 Magnet Avgerage Current Measurement for Main : Imag = -0.0020 +- 0.0000 A 17:32:08 Hall Probe Field Measurement, B = -0.000049 T 17:32:08 Hall Probe Field Measurement, B = -0.000050 T 17:32:09 Hall Probe Field Measurement, B = -0.000050 T 17:32:09 Hall Probe Field Measurement, B = -0.000050 T 17:32:09 Hall Probe Field Measurement, B = -0.000050 T 17:32:10 Hall Probe Field Measurement, B = -0.000049 T 17:32:10 Hall Probe Field Measurement: B = -0.000050 +/- 0.000000 T 17:32:10 Move Stage, position = 0.000000 17:32:10 Magnet system Main current ramping to 0.5000 A... 17:32:21 Move Stage, position = 0.000000 17:32:23 Stage Position Measurement, pos = 0.0000000 m 17:32:23 Magnet Current for Main, Imag = 0.4968 A Vtrans = 0.099301 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000286 A 17:32:24 Magnet Current for Main, Imag = 0.4968 A Vtrans = 0.099299 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000286 A 17:32:25 Magnet Current for Main, Imag = 0.4968 A Vtrans = 0.099301 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000286 A 17:32:26 Magnet Current for Main, Imag = 0.4968 A Vtrans = 0.099301 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000286 A 17:32:26 Magnet Avgerage Current Measurement for Main : Imag = 0.4968 +- 0.0000 A 17:32:26 Hall Probe Field Measurement, B = 0.004657 T 17:32:26 Hall Probe Field Measurement, B = 0.004657 T 17:32:27 Hall Probe Field Measurement, B = 0.004657 T 17:32:27 Hall Probe Field Measurement, B = 0.004657 T 17:32:27 Hall Probe Field Measurement, B = 0.004657 T 17:32:28 Hall Probe Field Measurement, B = 0.004656 T 17:32:28 Hall Probe Field Measurement: B = 0.004657 +/- 0.000000 T 17:32:28 Move Stage, position = 0.000000 17:32:28 Magnet system Main current ramping to 1.0000 A... 17:32:39 Move Stage, position = 0.000000 17:32:41 Stage Position Measurement, pos = 0.0000000 m 17:32:42 Magnet Current for Main, Imag = 0.9970 A Vtrans = 0.199345 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000286 A 17:32:42 Magnet Current for Main, Imag = 0.9970 A Vtrans = 0.199344 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000286 A 17:32:43 Magnet Current for Main, Imag = 0.9970 A Vtrans = 0.199343 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000286 A 17:32:44 Magnet Current for Main, Imag = 0.9970 A Vtrans = 0.199345 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000286 A 17:32:44 Magnet Avgerage Current Measurement for Main : Imag = 0.9970 +- 0.0000 A 17:32:44 Hall Probe Field Measurement, B = 0.009393 T 17:32:44 Hall Probe Field Measurement, B = 0.009393 T 17:32:45 Hall Probe Field Measurement, B = 0.009393 T 17:32:45 Hall Probe Field Measurement, B = 0.009393 T 17:32:45 Hall Probe Field Measurement, B = 0.009393 T 17:32:46 Hall Probe Field Measurement, B = 0.009394 T 17:32:46 Hall Probe Field Measurement: B = 0.009393 +/- 0.000000 T 17:32:46 Move Stage, position = 0.000000 17:32:46 Magnet system Main current ramping to 1.5000 A... 17:32:57 Move Stage, position = 0.000000 17:32:59 Stage Position Measurement, pos = 0.0000000 m 17:33:00 Magnet Current for Main, Imag = 1.5000 A Vtrans = 0.299945 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000286 A 17:33:00 Magnet Current for Main, Imag = 1.5000 A Vtrans = 0.299944 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000286 A 17:33:01 Magnet Current for Main, Imag = 1.5000 A Vtrans = 0.299945 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000286 A 17:33:02 Magnet Current for Main, Imag = 1.5000 A Vtrans = 0.299944 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000286 A 17:33:02 Magnet Avgerage Current Measurement for Main : Imag = 1.5000 +- 0.0000 A 17:33:02 Hall Probe Field Measurement, B = 0.014165 T 17:33:02 Hall Probe Field Measurement, B = 0.014165 T 17:33:03 Hall Probe Field Measurement, B = 0.014164 T 17:33:03 Hall Probe Field Measurement, B = 0.014164 T 17:33:03 Hall Probe Field Measurement, B = 0.014164 T 17:33:04 Hall Probe Field Measurement, B = 0.014164 T 17:33:04 Hall Probe Field Measurement: B = 0.014164 +/- 0.000000 T 17:33:04 Move Stage, position = 0.000000 17:33:04 Magnet system Main current ramping to 2.0000 A... 17:33:15 Move Stage, position = 0.000000 17:33:17 Stage Position Measurement, pos = 0.0000000 m 17:33:18 Magnet Current for Main, Imag = 1.9947 A Vtrans = 0.398891 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000286 A 17:33:18 Magnet Current for Main, Imag = 1.9947 A Vtrans = 0.398889 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000286 A 17:33:19 Magnet Current for Main, Imag = 1.9947 A Vtrans = 0.398890 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000286 A 17:33:20 Magnet Current for Main, Imag = 1.9947 A Vtrans = 0.398890 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000286 A 17:33:20 Magnet Avgerage Current Measurement for Main : Imag = 1.9947 +- 0.0000 A 17:33:20 Hall Probe Field Measurement, B = 0.018855 T 17:33:20 Hall Probe Field Measurement, B = 0.018855 T 17:33:21 Hall Probe Field Measurement, B = 0.018855 T 17:33:21 Hall Probe Field Measurement, B = 0.018855 T 17:33:21 Hall Probe Field Measurement, B = 0.018855 T 17:33:22 Hall Probe Field Measurement, B = 0.018855 T 17:33:22 Hall Probe Field Measurement: B = 0.018855 +/- 0.000000 T 17:33:22 Move Stage, position = 0.000000 17:33:22 Magnet system Main current ramping to 2.5000 A... 17:33:33 Move Stage, position = 0.000000 17:33:35 Stage Position Measurement, pos = 0.0000000 m 17:33:36 Magnet Current for Main, Imag = 2.4997 A Vtrans = 0.499885 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000286 A 17:33:36 Magnet Current for Main, Imag = 2.4997 A Vtrans = 0.499883 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000286 A 17:33:37 Magnet Current for Main, Imag = 2.4997 A Vtrans = 0.499884 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000286 A 17:33:38 Magnet Current for Main, Imag = 2.4997 A Vtrans = 0.499884 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000286 A 17:33:38 Magnet Avgerage Current Measurement for Main : Imag = 2.4997 +- 0.0000 A 17:33:38 Hall Probe Field Measurement, B = 0.023648 T 17:33:38 Hall Probe Field Measurement, B = 0.023648 T 17:33:39 Hall Probe Field Measurement, B = 0.023648 T 17:33:39 Hall Probe Field Measurement, B = 0.023648 T 17:33:39 Hall Probe Field Measurement, B = 0.023648 T 17:33:40 Hall Probe Field Measurement, B = 0.023649 T 17:33:40 Hall Probe Field Measurement: B = 0.023648 +/- 0.000000 T 17:33:40 Move Stage, position = 0.000000 17:33:40 Magnet system Main current ramping to 3.0000 A... 17:33:51 Move Stage, position = 0.000000 17:33:53 Stage Position Measurement, pos = 0.0000000 m 17:33:54 Magnet Current for Main, Imag = 3.0019 A Vtrans = 0.600319 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000286 A 17:33:54 Magnet Current for Main, Imag = 3.0019 A Vtrans = 0.600317 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000286 A 17:33:55 Magnet Current for Main, Imag = 3.0019 A Vtrans = 0.600320 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000286 A 17:33:56 Magnet Current for Main, Imag = 3.0019 A Vtrans = 0.600318 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000286 A 17:33:56 Magnet Avgerage Current Measurement for Main : Imag = 3.0019 +- 0.0000 A 17:33:56 Hall Probe Field Measurement, B = 0.028406 T 17:33:56 Hall Probe Field Measurement, B = 0.028406 T 17:33:57 Hall Probe Field Measurement, B = 0.028405 T 17:33:57 Hall Probe Field Measurement, B = 0.028405 T 17:33:58 Hall Probe Field Measurement, B = 0.028405 T 17:33:58 Hall Probe Field Measurement, B = 0.028405 T 17:33:59 Hall Probe Field Measurement: B = 0.028405 +/- 0.000000 T 17:33:59 Move Stage, position = 0.000000 17:33:59 Magnet system Main current ramping to 3.5000 A... 17:34:09 Move Stage, position = 0.000000 17:34:11 Stage Position Measurement, pos = 0.0000000 m 17:34:12 Magnet Current for Main, Imag = 3.4991 A Vtrans = 0.699761 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000286 A 17:34:13 Magnet Current for Main, Imag = 3.4991 A Vtrans = 0.699761 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000286 A 17:34:13 Magnet Current for Main, Imag = 3.4991 A Vtrans = 0.699760 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000286 A 17:34:14 Magnet Current for Main, Imag = 3.4991 A Vtrans = 0.699762 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000286 A 17:34:14 Magnet Avgerage Current Measurement for Main : Imag = 3.4991 +- 0.0000 A 17:34:14 Hall Probe Field Measurement, B = 0.033108 T 17:34:14 Hall Probe Field Measurement, B = 0.033108 T 17:34:15 Hall Probe Field Measurement, B = 0.033107 T 17:34:15 Hall Probe Field Measurement, B = 0.033107 T 17:34:16 Hall Probe Field Measurement, B = 0.033107 T 17:34:16 Hall Probe Field Measurement, B = 0.033107 T 17:34:17 Hall Probe Field Measurement: B = 0.033107 +/- 0.000000 T 17:34:17 Move Stage, position = 0.000000 17:34:17 Magnet system Main current ramping to 4.0000 A... 17:34:27 Move Stage, position = 0.000000 17:34:29 Stage Position Measurement, pos = 0.0000000 m 17:34:30 Magnet Current for Main, Imag = 3.9980 A Vtrans = 0.799543 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000286 A 17:34:31 Magnet Current for Main, Imag = 3.9980 A Vtrans = 0.799547 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000286 A 17:34:31 Magnet Current for Main, Imag = 3.9980 A Vtrans = 0.799546 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000286 A 17:34:32 Magnet Current for Main, Imag = 3.9980 A Vtrans = 0.799543 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000286 A 17:34:32 Magnet Avgerage Current Measurement for Main : Imag = 3.9980 +- 0.0000 A 17:34:32 Hall Probe Field Measurement, B = 0.037828 T 17:34:32 Hall Probe Field Measurement, B = 0.037828 T 17:34:33 Hall Probe Field Measurement, B = 0.037829 T 17:34:33 Hall Probe Field Measurement, B = 0.037828 T 17:34:34 Hall Probe Field Measurement, B = 0.037828 T 17:34:34 Hall Probe Field Measurement, B = 0.037828 T 17:34:35 Hall Probe Field Measurement: B = 0.037828 +/- 0.000000 T 17:34:35 Move Stage, position = 0.000000 17:34:35 Magnet system Main current ramping to 4.5000 A... 17:34:45 Move Stage, position = 0.000000 17:34:47 Stage Position Measurement, pos = 0.0000000 m 17:34:48 Magnet Current for Main, Imag = 4.4993 A Vtrans = 0.899798 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000286 A 17:34:49 Magnet Current for Main, Imag = 4.4993 A Vtrans = 0.899798 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000286 A 17:34:49 Magnet Current for Main, Imag = 4.4993 A Vtrans = 0.899799 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000286 A 17:34:50 Magnet Current for Main, Imag = 4.4993 A Vtrans = 0.899798 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000286 A 17:34:50 Magnet Avgerage Current Measurement for Main : Imag = 4.4993 +- 0.0000 A 17:34:50 Hall Probe Field Measurement, B = 0.042572 T 17:34:50 Hall Probe Field Measurement, B = 0.042572 T 17:34:51 Hall Probe Field Measurement, B = 0.042572 T 17:34:51 Hall Probe Field Measurement, B = 0.042572 T 17:34:52 Hall Probe Field Measurement, B = 0.042572 T 17:34:52 Hall Probe Field Measurement, B = 0.042572 T 17:34:53 Hall Probe Field Measurement: B = 0.042572 +/- 0.000000 T 17:34:53 Move Stage, position = 0.000000 17:34:53 Magnet system Main current ramping to 5.0000 A... 17:35:03 Move Stage, position = 0.000000 17:35:05 Stage Position Measurement, pos = 0.0000000 m 17:35:06 Magnet Current for Main, Imag = 5.0017 A Vtrans = 1.000279 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000286 A 17:35:07 Magnet Current for Main, Imag = 5.0017 A Vtrans = 1.000278 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000286 A 17:35:07 Magnet Current for Main, Imag = 5.0017 A Vtrans = 1.000279 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000286 A 17:35:08 Magnet Current for Main, Imag = 5.0017 A Vtrans = 1.000280 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000286 A 17:35:08 Magnet Avgerage Current Measurement for Main : Imag = 5.0017 +- 0.0000 A 17:35:08 Hall Probe Field Measurement, B = 0.047335 T 17:35:08 Hall Probe Field Measurement, B = 0.047336 T 17:35:09 Hall Probe Field Measurement, B = 0.047337 T 17:35:09 Hall Probe Field Measurement, B = 0.047338 T 17:35:10 Hall Probe Field Measurement, B = 0.047339 T 17:35:10 Hall Probe Field Measurement, B = 0.047339 T 17:35:11 Hall Probe Field Measurement: B = 0.047337 +/- 0.000001 T 17:35:11 Move Stage, position = 0.000000 17:35:11 Magnet system Main current ramping to 5.5000 A... 17:35:21 Move Stage, position = 0.000000 17:35:23 Stage Position Measurement, pos = 0.0000000 m 17:35:24 Magnet Current for Main, Imag = 5.4998 A Vtrans = 1.099903 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000286 A 17:35:25 Magnet Current for Main, Imag = 5.4998 A Vtrans = 1.099905 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000286 A 17:35:25 Magnet Current for Main, Imag = 5.4998 A Vtrans = 1.099905 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000286 A 17:35:26 Magnet Current for Main, Imag = 5.4998 A Vtrans = 1.099906 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000286 A 17:35:26 Magnet Avgerage Current Measurement for Main : Imag = 5.4998 +- 0.0000 A 17:35:26 Hall Probe Field Measurement, B = 0.052052 T 17:35:26 Hall Probe Field Measurement, B = 0.052053 T 17:35:27 Hall Probe Field Measurement, B = 0.052053 T 17:35:27 Hall Probe Field Measurement, B = 0.052053 T 17:35:28 Hall Probe Field Measurement, B = 0.052053 T 17:35:28 Hall Probe Field Measurement, B = 0.052053 T 17:35:29 Hall Probe Field Measurement: B = 0.052053 +/- 0.000000 T 17:35:29 Move Stage, position = 0.000000 17:35:29 Magnet system Main current ramping to 6.0000 A... 17:35:39 Move Stage, position = 0.000000 17:35:41 Stage Position Measurement, pos = 0.0000000 m 17:35:42 Magnet Current for Main, Imag = 6.0017 A Vtrans = 1.200288 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000286 A 17:35:43 Magnet Current for Main, Imag = 6.0017 A Vtrans = 1.200289 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000286 A 17:35:43 Magnet Current for Main, Imag = 6.0018 A Vtrans = 1.200293 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000286 A 17:35:44 Magnet Current for Main, Imag = 6.0018 A Vtrans = 1.200294 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000286 A 17:35:44 Magnet Avgerage Current Measurement for Main : Imag = 6.0017 +- 0.0000 A 17:35:44 Hall Probe Field Measurement, B = 0.056807 T 17:35:44 Hall Probe Field Measurement, B = 0.056806 T 17:35:45 Hall Probe Field Measurement, B = 0.056807 T 17:35:45 Hall Probe Field Measurement, B = 0.056807 T 17:35:46 Hall Probe Field Measurement, B = 0.056807 T 17:35:46 Hall Probe Field Measurement, B = 0.056807 T 17:35:47 Hall Probe Field Measurement: B = 0.056807 +/- 0.000000 T 17:35:47 Move Stage, position = 0.000000 17:35:47 Magnet system Main current ramping to 6.5000 A... 17:35:57 Move Stage, position = 0.000000 17:35:59 Stage Position Measurement, pos = 0.0000000 m 17:36:00 Magnet Current for Main, Imag = 6.4981 A Vtrans = 1.299560 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000286 A 17:36:01 Magnet Current for Main, Imag = 6.4981 A Vtrans = 1.299560 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000286 A 17:36:01 Magnet Current for Main, Imag = 6.4981 A Vtrans = 1.299565 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000286 A 17:36:02 Magnet Current for Main, Imag = 6.4981 A Vtrans = 1.299562 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000286 A 17:36:02 Magnet Avgerage Current Measurement for Main : Imag = 6.4981 +- 0.0000 A 17:36:02 Hall Probe Field Measurement, B = 0.061511 T 17:36:02 Hall Probe Field Measurement, B = 0.061511 T 17:36:03 Hall Probe Field Measurement, B = 0.061511 T 17:36:03 Hall Probe Field Measurement, B = 0.061511 T 17:36:04 Hall Probe Field Measurement, B = 0.061511 T 17:36:04 Hall Probe Field Measurement, B = 0.061511 T 17:36:05 Hall Probe Field Measurement: B = 0.061511 +/- 0.000000 T 17:36:05 Move Stage, position = 0.000000 17:36:05 Magnet system Main current ramping to 7.0000 A... 17:36:15 Move Stage, position = 0.000000 17:36:17 Stage Position Measurement, pos = 0.0000000 m 17:36:18 Magnet Current for Main, Imag = 6.9966 A Vtrans = 1.399266 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000286 A 17:36:19 Magnet Current for Main, Imag = 6.9966 A Vtrans = 1.399260 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000286 A 17:36:19 Magnet Current for Main, Imag = 6.9966 A Vtrans = 1.399257 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000286 A 17:36:20 Magnet Current for Main, Imag = 6.9966 A Vtrans = 1.399260 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000286 A 17:36:20 Magnet Avgerage Current Measurement for Main : Imag = 6.9966 +- 0.0000 A 17:36:20 Hall Probe Field Measurement, B = 0.066236 T 17:36:20 Hall Probe Field Measurement, B = 0.066237 T 17:36:21 Hall Probe Field Measurement, B = 0.066237 T 17:36:21 Hall Probe Field Measurement, B = 0.066238 T 17:36:22 Hall Probe Field Measurement, B = 0.066238 T 17:36:22 Hall Probe Field Measurement, B = 0.066238 T 17:36:23 Hall Probe Field Measurement: B = 0.066237 +/- 0.000001 T 17:36:23 Move Stage, position = 0.000000 17:36:23 Magnet system Main current ramping to 7.5000 A... 17:36:33 Move Stage, position = 0.000000 17:36:35 Stage Position Measurement, pos = 0.0000000 m 17:36:36 Magnet Current for Main, Imag = 7.4996 A Vtrans = 1.499855 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000286 A 17:36:37 Magnet Current for Main, Imag = 7.4995 A Vtrans = 1.499845 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000286 A 17:36:37 Magnet Current for Main, Imag = 7.4995 A Vtrans = 1.499846 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000286 A 17:36:38 Magnet Current for Main, Imag = 7.4995 A Vtrans = 1.499847 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000286 A 17:36:38 Magnet Avgerage Current Measurement for Main : Imag = 7.4995 +- 0.0000 A 17:36:38 Hall Probe Field Measurement, B = 0.071005 T 17:36:38 Hall Probe Field Measurement, B = 0.071005 T 17:36:39 Hall Probe Field Measurement, B = 0.071005 T 17:36:39 Hall Probe Field Measurement, B = 0.071006 T 17:36:40 Hall Probe Field Measurement, B = 0.071006 T 17:36:40 Hall Probe Field Measurement, B = 0.071005 T 17:36:41 Hall Probe Field Measurement: B = 0.071005 +/- 0.000000 T 17:36:41 Move Stage, position = 0.000000 17:36:41 Magnet system Main current ramping to 8.0000 A... 17:36:51 Move Stage, position = 0.000000 17:36:53 Stage Position Measurement, pos = 0.0000000 m 17:36:54 Magnet Current for Main, Imag = 7.9995 A Vtrans = 1.599846 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000286 A 17:36:55 Magnet Current for Main, Imag = 7.9995 A Vtrans = 1.599843 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000286 A 17:36:55 Magnet Current for Main, Imag = 7.9995 A Vtrans = 1.599843 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000286 A 17:36:56 Magnet Current for Main, Imag = 7.9995 A Vtrans = 1.599844 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000286 A 17:36:56 Magnet Avgerage Current Measurement for Main : Imag = 7.9995 +- 0.0000 A 17:36:56 Hall Probe Field Measurement, B = 0.075747 T 17:36:57 Hall Probe Field Measurement, B = 0.075746 T 17:36:57 Hall Probe Field Measurement, B = 0.075745 T 17:36:57 Hall Probe Field Measurement, B = 0.075745 T 17:36:58 Hall Probe Field Measurement, B = 0.075745 T 17:36:58 Hall Probe Field Measurement, B = 0.075745 T 17:36:59 Hall Probe Field Measurement: B = 0.075746 +/- 0.000001 T 17:36:59 Move Stage, position = 0.000000 17:36:59 Magnet system Main current ramping to 8.5000 A... 17:37:09 Move Stage, position = 0.000000 17:37:11 Stage Position Measurement, pos = 0.0000000 m 17:37:12 Magnet Current for Main, Imag = 8.4988 A Vtrans = 1.699704 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000286 A 17:37:13 Magnet Current for Main, Imag = 8.4988 A Vtrans = 1.699706 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000286 A 17:37:13 Magnet Current for Main, Imag = 8.4988 A Vtrans = 1.699704 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000286 A 17:37:14 Magnet Current for Main, Imag = 8.4988 A Vtrans = 1.699703 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000286 A 17:37:14 Magnet Avgerage Current Measurement for Main : Imag = 8.4988 +- 0.0000 A 17:37:14 Hall Probe Field Measurement, B = 0.080480 T 17:37:15 Hall Probe Field Measurement, B = 0.080480 T 17:37:15 Hall Probe Field Measurement, B = 0.080480 T 17:37:15 Hall Probe Field Measurement, B = 0.080479 T 17:37:16 Hall Probe Field Measurement, B = 0.080480 T 17:37:16 Hall Probe Field Measurement, B = 0.080480 T 17:37:17 Hall Probe Field Measurement: B = 0.080480 +/- 0.000000 T 17:37:17 Move Stage, position = 0.000000 17:37:17 Magnet system Main current ramping to 9.0000 A... 17:37:27 Move Stage, position = 0.000000 17:37:29 Stage Position Measurement, pos = 0.0000000 m 17:37:30 Magnet Current for Main, Imag = 9.0007 A Vtrans = 1.800090 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000286 A 17:37:31 Magnet Current for Main, Imag = 9.0007 A Vtrans = 1.800091 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000286 A 17:37:31 Magnet Current for Main, Imag = 9.0007 A Vtrans = 1.800091 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000286 A 17:37:32 Magnet Current for Main, Imag = 9.0007 A Vtrans = 1.800090 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000286 A 17:37:32 Magnet Avgerage Current Measurement for Main : Imag = 9.0007 +- 0.0000 A 17:37:32 Hall Probe Field Measurement, B = 0.085234 T 17:37:33 Hall Probe Field Measurement, B = 0.085234 T 17:37:33 Hall Probe Field Measurement, B = 0.085234 T 17:37:33 Hall Probe Field Measurement, B = 0.085235 T 17:37:34 Hall Probe Field Measurement, B = 0.085235 T 17:37:34 Hall Probe Field Measurement, B = 0.085235 T 17:37:35 Hall Probe Field Measurement: B = 0.085234 +/- 0.000001 T 17:37:35 Move Stage, position = 0.000000 17:37:35 Magnet system Main current ramping to 9.5000 A... 17:37:45 Move Stage, position = 0.000000 17:37:47 Stage Position Measurement, pos = 0.0000000 m 17:37:48 Magnet Current for Main, Imag = 9.4968 A Vtrans = 1.899295 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000286 A 17:37:49 Magnet Current for Main, Imag = 9.4968 A Vtrans = 1.899300 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000286 A 17:37:50 Magnet Current for Main, Imag = 9.4968 A Vtrans = 1.899303 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000286 A 17:37:50 Magnet Current for Main, Imag = 9.4968 A Vtrans = 1.899294 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000286 A 17:37:50 Magnet Avgerage Current Measurement for Main : Imag = 9.4968 +- 0.0000 A 17:37:50 Hall Probe Field Measurement, B = 0.089934 T 17:37:51 Hall Probe Field Measurement, B = 0.089935 T 17:37:51 Hall Probe Field Measurement, B = 0.089935 T 17:37:52 Hall Probe Field Measurement, B = 0.089935 T 17:37:52 Hall Probe Field Measurement, B = 0.089935 T 17:37:52 Hall Probe Field Measurement, B = 0.089935 T 17:37:53 Hall Probe Field Measurement: B = 0.089935 +/- 0.000000 T 17:37:53 Move Stage, position = 0.000000 17:37:53 Magnet system Main current ramping to 10.0000 A... 17:38:03 Move Stage, position = 0.000000 17:38:05 Stage Position Measurement, pos = 0.0000000 m 17:38:06 Magnet Current for Main, Imag = 10.0001 A Vtrans = 1.999956 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000286 A 17:38:07 Magnet Current for Main, Imag = 10.0001 A Vtrans = 1.999955 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000286 A 17:38:08 Magnet Current for Main, Imag = 10.0001 A Vtrans = 1.999953 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000286 A 17:38:08 Magnet Current for Main, Imag = 10.0001 A Vtrans = 1.999959 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000286 A 17:38:08 Magnet Avgerage Current Measurement for Main : Imag = 10.0001 +- 0.0000 A 17:38:08 Hall Probe Field Measurement, B = 0.094705 T 17:38:09 Hall Probe Field Measurement, B = 0.094704 T 17:38:09 Hall Probe Field Measurement, B = 0.094704 T 17:38:10 Hall Probe Field Measurement, B = 0.094704 T 17:38:10 Hall Probe Field Measurement, B = 0.094705 T 17:38:10 Hall Probe Field Measurement, B = 0.094705 T 17:38:11 Hall Probe Field Measurement: B = 0.094705 +/- 0.000001 T 17:38:11 Move Stage, position = 0.000000 17:38:11 Magnet system Main current ramping to 0.0000 A...