SLAC Magnetic Measurements Date: 10-06-2020 Time: 12:23:42 Project: LCLS-II Magnet Type: Radiation_Detection_System Magnet Name: SXU_004_Cell_27 Serial Number: PC33408 Run Number: 11 Measurement Device: Coil Operator: sda Comment: Int starting from 8.0008 (0.01%) with same 4 mm delta. 12:23:44 Ambient Measurement Device(s) Being Used, type = None 12:23:44 Magnet Voltage Measurement Device(s) Being Used, type = None 12:23:45 Temperature Measurement Device(s) Being Used, type = HPTHERM_HP34970 12:23:45 XMAG Device(s) Being Used, type = None 12:23:45 Ambient Measurement Device(s) Being Used, type = None 12:24:51   Voltage Measurement, V = 0.000000 V 12:24:52 Temperature Measurement, Probe Number = 0, T = 20.627000 C 12:24:52 Temperature Measurement, Probe Number = 1, T = 21.506000 C 12:24:52 Magnet Test Parameter Measurements P0 = Main (A) , Value = 0.00 P1 = Walker_Meas_Junction, Value = 20.63 P2 = 3458_Meas_Junction, Value = 21.51 12:24:53 Integrated Voltage Measurement... 12:25:39 Voltage Offset Correction: v_off_ini = -0.000001526 V, v_off_fin = 0.000001255 V 12:25:39 Integrated Voltage Measurement Delta Gap = 0.0040013 m VT = 0.143534750 Vs 12:25:39 Integrated Voltage Measurement... 12:26:22 Voltage Offset Correction: v_off_ini = -0.000000189 V, v_off_fin = -0.000002158 V 12:26:22 Integrated Voltage Measurement Delta Gap = -0.0040007 m VT = -0.143484920 Vs 12:26:23   Voltage Measurement, V = 0.000000 V 12:26:24 Temperature Measurement, Probe Number = 0, T = 20.587000 C 12:26:24 Temperature Measurement, Probe Number = 1, T = 21.498000 C 12:26:24 Magnet Test Parameter Measurements P0 = Main (A) , Value = 0.00 P1 = Walker_Meas_Junction, Value = 20.59 P2 = 3458_Meas_Junction, Value = 21.50 12:26:24 Integrated Voltage Measurement... 12:27:03 Voltage Offset Correction: v_off_ini = -0.000000943 V, v_off_fin = 0.000000136 V 12:27:03 Integrated Voltage Measurement Delta Gap = 0.0040004 m VT = 0.143498317 Vs 12:27:03 Integrated Voltage Measurement... 12:27:46 Voltage Offset Correction: v_off_ini = 0.000000046 V, v_off_fin = 0.000000187 V 12:27:46 Integrated Voltage Measurement Delta Gap = -0.0040008 m VT = -0.143499092 Vs 12:27:47   Voltage Measurement, V = 0.000000 V 12:27:47 Temperature Measurement, Probe Number = 0, T = 20.607000 C 12:27:48 Temperature Measurement, Probe Number = 1, T = 21.489000 C 12:27:48 Magnet Test Parameter Measurements P0 = Main (A) , Value = 0.00 P1 = Walker_Meas_Junction, Value = 20.61 P2 = 3458_Meas_Junction, Value = 21.49 12:27:48 Integrated Voltage Measurement... 12:28:24 Voltage Offset Correction: v_off_ini = -0.000000699 V, v_off_fin = 0.000001814 V 12:28:24 Integrated Voltage Measurement Delta Gap = 0.0040003 m VT = 0.143487005 Vs 12:28:24 Integrated Voltage Measurement... 12:28:58 Voltage Offset Correction: v_off_ini = -0.000000127 V, v_off_fin = -0.000003080 V 12:28:58 Integrated Voltage Measurement Delta Gap = -0.0040007 m VT = -0.143492432 Vs 12:28:59   Voltage Measurement, V = 0.000000 V 12:29:00 Temperature Measurement, Probe Number = 0, T = 20.654000 C 12:29:01 Temperature Measurement, Probe Number = 1, T = 21.498000 C 12:29:01 Magnet Test Parameter Measurements P0 = Main (A) , Value = 0.00 P1 = Walker_Meas_Junction, Value = 20.65 P2 = 3458_Meas_Junction, Value = 21.50 12:29:01 Integrated Voltage Measurement... 12:29:37 Voltage Offset Correction: v_off_ini = -0.000001293 V, v_off_fin = 0.000000641 V 12:29:37 Integrated Voltage Measurement Delta Gap = 0.0040008 m VT = 0.143513585 Vs 12:29:37 Integrated Voltage Measurement... 12:30:13 Voltage Offset Correction: v_off_ini = -0.000000058 V, v_off_fin = -0.000003012 V 12:30:13 Integrated Voltage Measurement Delta Gap = -0.0040008 m VT = -0.143484455 Vs 12:30:14   Voltage Measurement, V = 0.000000 V 12:30:15 Temperature Measurement, Probe Number = 0, T = 20.683000 C 12:30:15 Temperature Measurement, Probe Number = 1, T = 21.505000 C 12:30:15 Magnet Test Parameter Measurements P0 = Main (A) , Value = 0.00 P1 = Walker_Meas_Junction, Value = 20.68 P2 = 3458_Meas_Junction, Value = 21.50 12:30:15 Integrated Voltage Measurement... 12:30:51 Voltage Offset Correction: v_off_ini = -0.000001297 V, v_off_fin = 0.000000364 V 12:30:51 Integrated Voltage Measurement Delta Gap = 0.0040009 m VT = 0.143522902 Vs 12:30:51 Integrated Voltage Measurement... 12:31:29 Voltage Offset Correction: v_off_ini = 0.000000067 V, v_off_fin = -0.000002315 V 12:31:29 Integrated Voltage Measurement Delta Gap = -0.0040007 m VT = -0.143485802 Vs 12:31:30   Voltage Measurement, V = 0.000000 V 12:31:31 Temperature Measurement, Probe Number = 0, T = 20.706000 C 12:31:32 Temperature Measurement, Probe Number = 1, T = 21.511000 C 12:31:32 Magnet Test Parameter Measurements P0 = Main (A) , Value = 0.00 P1 = Walker_Meas_Junction, Value = 20.71 P2 = 3458_Meas_Junction, Value = 21.51 12:31:32 Integrated Voltage Measurement... 12:32:06 Voltage Offset Correction: v_off_ini = -0.000001281 V, v_off_fin = 0.000000583 V 12:32:06 Integrated Voltage Measurement Delta Gap = 0.0040006 m VT = 0.143505339 Vs 12:32:07 Integrated Voltage Measurement... 12:32:43 Voltage Offset Correction: v_off_ini = 0.000000043 V, v_off_fin = -0.000002140 V 12:32:43 Integrated Voltage Measurement Delta Gap = -0.0040008 m VT = -0.143487258 Vs 12:32:44   Voltage Measurement, V = 0.000000 V 12:32:45 Temperature Measurement, Probe Number = 0, T = 20.720000 C 12:32:45 Temperature Measurement, Probe Number = 1, T = 21.508000 C 12:32:45 Magnet Test Parameter Measurements P0 = Main (A) , Value = 0.00 P1 = Walker_Meas_Junction, Value = 20.72 P2 = 3458_Meas_Junction, Value = 21.51 12:32:45 Integrated Voltage Measurement... 12:33:21 Voltage Offset Correction: v_off_ini = -0.000001261 V, v_off_fin = 0.000000530 V 12:33:21 Integrated Voltage Measurement Delta Gap = 0.0040007 m VT = 0.143510657 Vs 12:33:21 Integrated Voltage Measurement... 12:33:55 Voltage Offset Correction: v_off_ini = 0.000000064 V, v_off_fin = -0.000001734 V 12:33:55 Integrated Voltage Measurement Delta Gap = -0.0040008 m VT = -0.143493845 Vs 12:33:56   Voltage Measurement, V = 0.000000 V 12:33:57 Temperature Measurement, Probe Number = 0, T = 20.735000 C 12:33:58 Temperature Measurement, Probe Number = 1, T = 21.504000 C 12:33:58 Magnet Test Parameter Measurements P0 = Main (A) , Value = 0.00 P1 = Walker_Meas_Junction, Value = 20.73 P2 = 3458_Meas_Junction, Value = 21.50 12:33:58 Integrated Voltage Measurement... 12:35:02 Voltage Offset Correction: v_off_ini = -0.000000819 V, v_off_fin = 0.000001195 V 12:35:02 Integrated Voltage Measurement Delta Gap = 0.0040010 m VT = 0.143517878 Vs 12:35:02 Integrated Voltage Measurement... 12:35:36 Voltage Offset Correction: v_off_ini = -0.000000126 V, v_off_fin = -0.000002102 V 12:35:36 Integrated Voltage Measurement Delta Gap = -0.0040007 m VT = -0.143480955 Vs 12:35:37   Voltage Measurement, V = 0.000000 V 12:35:37 Temperature Measurement, Probe Number = 0, T = 20.791000 C 12:35:38 Temperature Measurement, Probe Number = 1, T = 21.496000 C 12:35:38 Magnet Test Parameter Measurements P0 = Main (A) , Value = 0.00 P1 = Walker_Meas_Junction, Value = 20.79 P2 = 3458_Meas_Junction, Value = 21.50 12:35:38 Integrated Voltage Measurement... 12:36:14 Voltage Offset Correction: v_off_ini = -0.000001238 V, v_off_fin = 0.000000399 V 12:36:14 Integrated Voltage Measurement Delta Gap = 0.0040006 m VT = 0.143502838 Vs 12:36:14 Integrated Voltage Measurement... 12:36:48 Voltage Offset Correction: v_off_ini = -0.000000001 V, v_off_fin = -0.000002886 V 12:36:48 Integrated Voltage Measurement Delta Gap = -0.0040007 m VT = -0.143495120 Vs 12:36:49   Voltage Measurement, V = 0.000000 V 12:36:50 Temperature Measurement, Probe Number = 0, T = 20.788000 C 12:36:51 Temperature Measurement, Probe Number = 1, T = 21.499000 C 12:36:51 Magnet Test Parameter Measurements P0 = Main (A) , Value = 0.00 P1 = Walker_Meas_Junction, Value = 20.79 P2 = 3458_Meas_Junction, Value = 21.50 12:36:51 Integrated Voltage Measurement... 12:37:26 Voltage Offset Correction: v_off_ini = -0.000001374 V, v_off_fin = -0.000000064 V 12:37:26 Integrated Voltage Measurement Delta Gap = 0.0040009 m VT = 0.143526012 Vs 12:37:26 Integrated Voltage Measurement... 12:38:04 Voltage Offset Correction: v_off_ini = 0.000000050 V, v_off_fin = -0.000002822 V 12:38:04 Integrated Voltage Measurement Delta Gap = -0.0040008 m VT = -0.143488443 Vs