SLAC Magnetic Measurements Date: 05-29-2019 Time: 08:59:15 Project: LCLS-II Magnet Type: Dipole Magnet Name: 4509 Serial Number: Run Number: 2 Measurement Device: 10wire Operator: sda Comment: Integrated field vs trim current with main at nominal current of 126.4349. Standardization for MAIN: 3 cycles from 0.0 A to 200.0 A to 0.0 A Test Currents for MAIN (A): 126.43 126.43 126.43 126.43 126.43 126.43 126.43 Test Currents for TRIM (A): 0.00 -6.00 0.00 6.00 0.00 -6.00 6.00 08:59:43 Position moved: 0.000000 m 08:59:58 Magnet Current for MAIN, Imag = -0.0047 A Vtrans = -0.000047 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 08:59:59 Magnet Current for MAIN, Imag = -0.0047 A Vtrans = -0.000047 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 09:00:00 Magnet Current for MAIN, Imag = -0.0047 A Vtrans = -0.000047 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 09:00:01 Magnet Current for MAIN, Imag = -0.0047 A Vtrans = -0.000047 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 09:00:01 Magnet Avgerage Current Measurement for MAIN : Imag = -0.0047 +- 0.0000 A 09:00:01 Initial Zero Current Measurement for MAIN, Imag_zero = -0.00468863 A 09:00:02 Magnet Current for TRIM, Imag = 0.0011 A Vtrans = 0.000093 V, Factor V/A = 0.083333 V/A, Zero Current = 0.000000 A 09:00:03 Magnet Current for TRIM, Imag = 0.0011 A Vtrans = 0.000093 V, Factor V/A = 0.083333 V/A, Zero Current = 0.000000 A 09:00:03 Magnet Current for TRIM, Imag = 0.0011 A Vtrans = 0.000093 V, Factor V/A = 0.083333 V/A, Zero Current = 0.000000 A 09:00:04 Magnet Current for TRIM, Imag = 0.0011 A Vtrans = 0.000093 V, Factor V/A = 0.083333 V/A, Zero Current = 0.000000 A 09:00:04 Magnet Avgerage Current Measurement for TRIM : Imag = 0.0011 +- 0.0000 A 09:00:55 Initial Zero Current Measurement for TRIM, Imag_zero = 0.00111608 A 09:00:55 Ambient Measurement Device(s) Being Used, type = None 09:00:56 Magnet Voltage Measurement Device(s) Being Used, type = HP34970 09:00:56 Temperature Measurement Device(s) Being Used, type = HPTHERM_HP34970 09:00:56 XMAG Device(s) Being Used, type = None 09:00:57 Ambient Measurement Device(s) Being Used, type = HP34970 09:00:57 Magnet system MAIN current ramping to 200.0000 A... 09:02:59 Magnet system MAIN current ramping to 0.0000 A... 09:05:01 Magnet system MAIN current ramping to 200.0000 A... 09:07:04 Magnet system MAIN current ramping to 0.0000 A... 09:09:06 Magnet system MAIN current ramping to 200.0000 A... 09:11:08 Magnet system MAIN current ramping to 0.0000 A... 09:13:13 Position moved: 0.000000 m 09:13:16 Magnet system MAIN current ramping to 126.4349 A... 09:14:37 Magnet system TRIM current ramping to 0.0000 A... 09:14:48 Magnet Current for MAIN, Imag = 126.3325 A Vtrans = 1.263278 V, Factor V/A = 0.010000 V/A, Zero Current = -0.004689 A 09:14:49 Magnet Current for TRIM, Imag = 0.0002 A Vtrans = 0.000110 V, Factor V/A = 0.083333 V/A, Zero Current = 0.001116 A 09:14:49 MAIN (V) Voltage Measurement, V = 8.976441 V 09:14:50 Temperature Measurement, Probe Number = 0, T = 27.421000 C 09:14:51 Temperature Measurement, Probe Number = 1, T = 25.846000 C 09:14:51 Temperature Measurement, Probe Number = 2, T = 19.817000 C 09:14:52 Sensor Voltage Measurement, Sensor Number = 0, V = 2.320582 V 09:14:52 Sensor Measurement, Sensor Number = 0, Sensor Type = WATER_HPLEIGHTON_T Raw Sensor Output = 2.320582 (V, A, ohm, ...), Measured Value = 28.341419 (C, atm, Hz, ...) 09:14:53 Sensor Voltage Measurement, Sensor Number = 1, V = 2.298080 V 09:14:53 Sensor Measurement, Sensor Number = 1, Sensor Type = WATER_HPLEIGHTON_T Raw Sensor Output = 2.298080 (V, A, ohm, ...), Measured Value = 28.028893 (C, atm, Hz, ...) 09:14:53 Sensor Voltage Measurement, Sensor Number = 2, V = 2.655642 V 09:14:53 Sensor Measurement, Sensor Number = 2, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 2.655642 (V, A, ohm, ...), Measured Value = 124.173120 (C, atm, Hz, ...) 09:14:54 Sensor Voltage Measurement, Sensor Number = 3, V = 1.093732 V 09:14:54 Sensor Measurement, Sensor Number = 3, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 1.093732 (V, A, ohm, ...), Measured Value = 7.029907 (C, atm, Hz, ...) 09:14:54 Magnet Test Parameter Measurements P0 = MAIN, Value = 126.33 P1 = TRIM, Value = 0.00 P2 = MAIN (V), Value = 8.98 P3 = Coil, Value = 27.42 P4 = Magnet Core, Value = 25.85 P5 = Ambient Temp (C), Value = 19.82 P6 = Supply Temp(C), Value = 28.34 P7 = Return Temp(C), Value = 28.03 P8 = Supply Pres(psi), Value = 124.17 P9 = Return Pres(psi), Value = 7.03 09:14:54 Magnetic Field Measurement Measuring the magnet strength... 09:14:54 Integrated Voltage Measurement... 09:14:55 Magnet Current for MAIN, Imag = 126.3329 A Vtrans = 1.263282 V, Factor V/A = 0.010000 V/A, Zero Current = -0.004689 A 09:14:56 Magnet Current for TRIM, Imag = 0.0003 A Vtrans = 0.000119 V, Factor V/A = 0.083333 V/A, Zero Current = 0.001116 A 09:15:05 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0090000 m VT = 0.058746439 Vs 09:15:13 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0090000 m VT = -0.058780667 Vs 09:15:13 Magnet Current for MAIN, Imag = 126.3324 A Vtrans = 1.263277 V, Factor V/A = 0.010000 V/A, Zero Current = -0.004689 A 09:15:14 Magnet Current for TRIM, Imag = 0.0001 A Vtrans = 0.000105 V, Factor V/A = 0.083333 V/A, Zero Current = 0.001116 A 09:15:22 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0090000 m VT = 0.058736740 Vs 09:15:30 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0090000 m VT = -0.058809183 Vs 09:15:30 Magnet Current for MAIN, Imag = 126.3325 A Vtrans = 1.263278 V, Factor V/A = 0.010000 V/A, Zero Current = -0.004689 A 09:15:31 Magnet Current for TRIM, Imag = 0.0002 A Vtrans = 0.000114 V, Factor V/A = 0.083333 V/A, Zero Current = 0.001116 A 09:15:39 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0090000 m VT = 0.058754394 Vs 09:15:47 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0090000 m VT = -0.058783690 Vs 09:15:47 Magnet Current for MAIN, Imag = 126.3324 A Vtrans = 1.263277 V, Factor V/A = 0.010000 V/A, Zero Current = -0.004689 A 09:15:48 Magnet Current for TRIM, Imag = 0.0003 A Vtrans = 0.000119 V, Factor V/A = 0.083333 V/A, Zero Current = 0.001116 A 09:15:56 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0090000 m VT = 0.058753735 Vs 09:16:04 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0090000 m VT = -0.058787913 Vs 09:16:04 Integrated Voltage Average and RMS of Measurements X0 = 0.0000000 m, Delta X = 0.0090000 m VT = 0.058769095 +- 0.000003488 Vs 09:16:04 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0000000 m, dx = 0.0090000 m, Nturns = 10.0 VT = 0.058769095 +- 0.000003488 Vs BL = 0.65298995 +- 0.00003875 Tm 09:16:08 Position moved: 0.000000 m 09:16:08 BL vs X Fit Coefficients: Cn sigCn n (Tm/m^n) ----- --------+-------- 0 0.65299 0.00004 Compute Fit Residuals: X0 BL sigBL BLfit BL-BLfit (m) (Tm) (Tm) (Tm) (Tm) ----------- --------+-------- -------- -------- 0.0000000 0.65299 0.00004 0.65299 0.00000 09:16:08 Magnet Strength Calculation Harmonic Number = 1 Number Of X0 Positions = 1 X0 BL sigBL (m) (Tm) (Tm) -------- --------+-------- 0.0000 0.65299 0.00004 SL = 0.6529899 +- 0.0000388 Tm^2/m^n 09:16:08 Magnet Strength Measurement Harmonic Number = 1 SL = 0.6529899 +- 0.0000388 Tm^2/m^n 09:16:11 Position moved: 0.000000 m 09:16:11 Magnet system MAIN current ramping to 126.4349 A... 09:16:21 Magnet system TRIM current ramping to -6.0000 A... 09:16:38 Magnet Current for MAIN, Imag = 126.3323 A Vtrans = 1.263276 V, Factor V/A = 0.010000 V/A, Zero Current = -0.004689 A 09:16:39 Magnet Current for TRIM, Imag = -6.0040 A Vtrans = -0.500237 V, Factor V/A = 0.083333 V/A, Zero Current = 0.001116 A 09:16:40 MAIN (V) Voltage Measurement, V = 8.985938 V 09:16:40 Temperature Measurement, Probe Number = 0, T = 27.464000 C 09:16:41 Temperature Measurement, Probe Number = 1, T = 25.808000 C 09:16:41 Temperature Measurement, Probe Number = 2, T = 19.793000 C 09:16:42 Sensor Voltage Measurement, Sensor Number = 0, V = 2.319804 V 09:16:42 Sensor Measurement, Sensor Number = 0, Sensor Type = WATER_HPLEIGHTON_T Raw Sensor Output = 2.319804 (V, A, ohm, ...), Measured Value = 28.330615 (C, atm, Hz, ...) 09:16:43 Sensor Voltage Measurement, Sensor Number = 1, V = 2.297609 V 09:16:43 Sensor Measurement, Sensor Number = 1, Sensor Type = WATER_HPLEIGHTON_T Raw Sensor Output = 2.297609 (V, A, ohm, ...), Measured Value = 28.022343 (C, atm, Hz, ...) 09:16:43 Sensor Voltage Measurement, Sensor Number = 2, V = 2.645624 V 09:16:44 Sensor Measurement, Sensor Number = 2, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 2.645624 (V, A, ohm, ...), Measured Value = 123.421815 (C, atm, Hz, ...) 09:16:44 Sensor Voltage Measurement, Sensor Number = 3, V = 1.093954 V 09:16:44 Sensor Measurement, Sensor Number = 3, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 1.093954 (V, A, ohm, ...), Measured Value = 7.046587 (C, atm, Hz, ...) 09:16:44 Magnet Test Parameter Measurements P0 = MAIN, Value = 126.33 P1 = TRIM, Value = -6.00 P2 = MAIN (V), Value = 8.99 P3 = Coil, Value = 27.46 P4 = Magnet Core, Value = 25.81 P5 = Ambient Temp (C), Value = 19.79 P6 = Supply Temp(C), Value = 28.33 P7 = Return Temp(C), Value = 28.02 P8 = Supply Pres(psi), Value = 123.42 P9 = Return Pres(psi), Value = 7.05 09:16:44 Magnetic Field Measurement Measuring the magnet strength... 09:16:44 Integrated Voltage Measurement... 09:16:45 Magnet Current for MAIN, Imag = 126.3324 A Vtrans = 1.263277 V, Factor V/A = 0.010000 V/A, Zero Current = -0.004689 A 09:16:46 Magnet Current for TRIM, Imag = -6.0040 A Vtrans = -0.500237 V, Factor V/A = 0.083333 V/A, Zero Current = 0.001116 A 09:16:55 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0090000 m VT = 0.056897095 Vs 09:17:03 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0090000 m VT = -0.056912265 Vs 09:17:03 Magnet Current for MAIN, Imag = 126.3321 A Vtrans = 1.263275 V, Factor V/A = 0.010000 V/A, Zero Current = -0.004689 A 09:17:04 Magnet Current for TRIM, Imag = -6.0039 A Vtrans = -0.500236 V, Factor V/A = 0.083333 V/A, Zero Current = 0.001116 A 09:17:12 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0090000 m VT = 0.056877980 Vs 09:17:20 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0090000 m VT = -0.056911445 Vs 09:17:20 Magnet Current for MAIN, Imag = 126.3316 A Vtrans = 1.263269 V, Factor V/A = 0.010000 V/A, Zero Current = -0.004689 A 09:17:21 Magnet Current for TRIM, Imag = -6.0039 A Vtrans = -0.500234 V, Factor V/A = 0.083333 V/A, Zero Current = 0.001116 A 09:17:29 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0090000 m VT = 0.056887491 Vs 09:17:37 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0090000 m VT = -0.056902313 Vs 09:17:38 Magnet Current for MAIN, Imag = 126.3316 A Vtrans = 1.263269 V, Factor V/A = 0.010000 V/A, Zero Current = -0.004689 A 09:17:38 Magnet Current for TRIM, Imag = -6.0039 A Vtrans = -0.500234 V, Factor V/A = 0.083333 V/A, Zero Current = 0.001116 A 09:17:46 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0090000 m VT = 0.056888035 Vs 09:17:54 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0090000 m VT = -0.056901506 Vs 09:17:54 Integrated Voltage Average and RMS of Measurements X0 = 0.0000000 m, Delta X = 0.0090000 m VT = 0.056897266 +- 0.000004281 Vs 09:17:54 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0000000 m, dx = 0.0090000 m, Nturns = 10.0 VT = 0.056897266 +- 0.000004281 Vs BL = 0.63219185 +- 0.00004757 Tm 09:17:58 Position moved: 0.000000 m 09:17:58 BL vs X Fit Coefficients: Cn sigCn n (Tm/m^n) ----- --------+-------- 0 0.63219 0.00005 Compute Fit Residuals: X0 BL sigBL BLfit BL-BLfit (m) (Tm) (Tm) (Tm) (Tm) ----------- --------+-------- -------- -------- 0.0000000 0.63219 0.00005 0.63219 0.00000 09:17:58 Magnet Strength Calculation Harmonic Number = 1 Number Of X0 Positions = 1 X0 BL sigBL (m) (Tm) (Tm) -------- --------+-------- 0.0000 0.63219 0.00005 SL = 0.6321918 +- 0.0000476 Tm^2/m^n 09:17:58 Magnet Strength Measurement Harmonic Number = 1 SL = 0.6321918 +- 0.0000476 Tm^2/m^n 09:18:01 Position moved: 0.000000 m 09:18:01 Magnet system MAIN current ramping to 126.4349 A... 09:18:11 Magnet system TRIM current ramping to 0.0000 A... 09:18:28 Magnet Current for MAIN, Imag = 126.3310 A Vtrans = 1.263263 V, Factor V/A = 0.010000 V/A, Zero Current = -0.004689 A 09:18:29 Magnet Current for TRIM, Imag = 0.0003 A Vtrans = 0.000116 V, Factor V/A = 0.083333 V/A, Zero Current = 0.001116 A 09:18:30 MAIN (V) Voltage Measurement, V = 8.988025 V 09:18:30 Temperature Measurement, Probe Number = 0, T = 27.499000 C 09:18:31 Temperature Measurement, Probe Number = 1, T = 25.867000 C 09:18:32 Temperature Measurement, Probe Number = 2, T = 19.771000 C 09:18:32 Sensor Voltage Measurement, Sensor Number = 0, V = 2.320550 V 09:18:32 Sensor Measurement, Sensor Number = 0, Sensor Type = WATER_HPLEIGHTON_T Raw Sensor Output = 2.320550 (V, A, ohm, ...), Measured Value = 28.340978 (C, atm, Hz, ...) 09:18:33 Sensor Voltage Measurement, Sensor Number = 1, V = 2.298997 V 09:18:33 Sensor Measurement, Sensor Number = 1, Sensor Type = WATER_HPLEIGHTON_T Raw Sensor Output = 2.298997 (V, A, ohm, ...), Measured Value = 28.041621 (C, atm, Hz, ...) 09:18:34 Sensor Voltage Measurement, Sensor Number = 2, V = 2.659605 V 09:18:34 Sensor Measurement, Sensor Number = 2, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 2.659605 (V, A, ohm, ...), Measured Value = 124.470360 (C, atm, Hz, ...) 09:18:34 Sensor Voltage Measurement, Sensor Number = 3, V = 1.093989 V 09:18:34 Sensor Measurement, Sensor Number = 3, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 1.093989 (V, A, ohm, ...), Measured Value = 7.049160 (C, atm, Hz, ...) 09:18:34 Magnet Test Parameter Measurements P0 = MAIN, Value = 126.33 P1 = TRIM, Value = 0.00 P2 = MAIN (V), Value = 8.99 P3 = Coil, Value = 27.50 P4 = Magnet Core, Value = 25.87 P5 = Ambient Temp (C), Value = 19.77 P6 = Supply Temp(C), Value = 28.34 P7 = Return Temp(C), Value = 28.04 P8 = Supply Pres(psi), Value = 124.47 P9 = Return Pres(psi), Value = 7.05 09:18:34 Magnetic Field Measurement Measuring the magnet strength... 09:18:34 Integrated Voltage Measurement... 09:18:35 Magnet Current for MAIN, Imag = 126.3310 A Vtrans = 1.263263 V, Factor V/A = 0.010000 V/A, Zero Current = -0.004689 A 09:18:36 Magnet Current for TRIM, Imag = 0.0002 A Vtrans = 0.000114 V, Factor V/A = 0.083333 V/A, Zero Current = 0.001116 A 09:18:45 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0090000 m VT = 0.058754718 Vs 09:18:53 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0090000 m VT = -0.058771327 Vs 09:18:53 Magnet Current for MAIN, Imag = 126.3309 A Vtrans = 1.263262 V, Factor V/A = 0.010000 V/A, Zero Current = -0.004689 A 09:18:54 Magnet Current for TRIM, Imag = 0.0002 A Vtrans = 0.000111 V, Factor V/A = 0.083333 V/A, Zero Current = 0.001116 A 09:19:02 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0090000 m VT = 0.058779196 Vs 09:19:10 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0090000 m VT = -0.058757153 Vs 09:19:10 Magnet Current for MAIN, Imag = 126.3311 A Vtrans = 1.263264 V, Factor V/A = 0.010000 V/A, Zero Current = -0.004689 A 09:19:11 Magnet Current for TRIM, Imag = 0.0002 A Vtrans = 0.000113 V, Factor V/A = 0.083333 V/A, Zero Current = 0.001116 A 09:19:19 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0090000 m VT = 0.058772657 Vs 09:19:27 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0090000 m VT = -0.058762047 Vs 09:19:28 Magnet Current for MAIN, Imag = 126.3309 A Vtrans = 1.263262 V, Factor V/A = 0.010000 V/A, Zero Current = -0.004689 A 09:19:28 Magnet Current for TRIM, Imag = 0.0003 A Vtrans = 0.000114 V, Factor V/A = 0.083333 V/A, Zero Current = 0.001116 A 09:19:36 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0090000 m VT = 0.058765459 Vs 09:19:44 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0090000 m VT = -0.058765221 Vs 09:19:44 Integrated Voltage Average and RMS of Measurements X0 = 0.0000000 m, Delta X = 0.0090000 m VT = 0.058765972 +- 0.000001991 Vs 09:19:44 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0000000 m, dx = 0.0090000 m, Nturns = 10.0 VT = 0.058765972 +- 0.000001991 Vs BL = 0.65295525 +- 0.00002212 Tm 09:19:48 Position moved: 0.000000 m 09:19:48 BL vs X Fit Coefficients: Cn sigCn n (Tm/m^n) ----- --------+-------- 0 0.65296 0.00002 Compute Fit Residuals: X0 BL sigBL BLfit BL-BLfit (m) (Tm) (Tm) (Tm) (Tm) ----------- --------+-------- -------- -------- 0.0000000 0.65296 0.00002 0.65296 0.00000 09:19:48 Magnet Strength Calculation Harmonic Number = 1 Number Of X0 Positions = 1 X0 BL sigBL (m) (Tm) (Tm) -------- --------+-------- 0.0000 0.65296 0.00002 SL = 0.6529552 +- 0.0000221 Tm^2/m^n 09:19:48 Magnet Strength Measurement Harmonic Number = 1 SL = 0.6529552 +- 0.0000221 Tm^2/m^n 09:19:51 Position moved: 0.000000 m 09:19:51 Magnet system MAIN current ramping to 126.4349 A... 09:20:02 Magnet system TRIM current ramping to 6.0000 A... 09:20:18 Magnet Current for MAIN, Imag = 126.3301 A Vtrans = 1.263254 V, Factor V/A = 0.010000 V/A, Zero Current = -0.004689 A 09:20:19 Magnet Current for TRIM, Imag = 6.0032 A Vtrans = 0.500362 V, Factor V/A = 0.083333 V/A, Zero Current = 0.001116 A 09:20:20 MAIN (V) Voltage Measurement, V = 8.988369 V 09:20:20 Temperature Measurement, Probe Number = 0, T = 27.516000 C 09:20:21 Temperature Measurement, Probe Number = 1, T = 25.898000 C 09:20:22 Temperature Measurement, Probe Number = 2, T = 19.790000 C 09:20:22 Sensor Voltage Measurement, Sensor Number = 0, V = 2.320226 V 09:20:22 Sensor Measurement, Sensor Number = 0, Sensor Type = WATER_HPLEIGHTON_T Raw Sensor Output = 2.320226 (V, A, ohm, ...), Measured Value = 28.336476 (C, atm, Hz, ...) 09:20:23 Sensor Voltage Measurement, Sensor Number = 1, V = 2.302956 V 09:20:23 Sensor Measurement, Sensor Number = 1, Sensor Type = WATER_HPLEIGHTON_T Raw Sensor Output = 2.302956 (V, A, ohm, ...), Measured Value = 28.096612 (C, atm, Hz, ...) 09:20:24 Sensor Voltage Measurement, Sensor Number = 2, V = 2.663653 V 09:20:24 Sensor Measurement, Sensor Number = 2, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 2.663653 (V, A, ohm, ...), Measured Value = 124.773982 (C, atm, Hz, ...) 09:20:24 Sensor Voltage Measurement, Sensor Number = 3, V = 1.093708 V 09:20:24 Sensor Measurement, Sensor Number = 3, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 1.093708 (V, A, ohm, ...), Measured Value = 7.028093 (C, atm, Hz, ...) 09:20:24 Magnet Test Parameter Measurements P0 = MAIN, Value = 126.33 P1 = TRIM, Value = 6.00 P2 = MAIN (V), Value = 8.99 P3 = Coil, Value = 27.52 P4 = Magnet Core, Value = 25.90 P5 = Ambient Temp (C), Value = 19.79 P6 = Supply Temp(C), Value = 28.34 P7 = Return Temp(C), Value = 28.10 P8 = Supply Pres(psi), Value = 124.77 P9 = Return Pres(psi), Value = 7.03 09:20:25 Magnetic Field Measurement Measuring the magnet strength... 09:20:25 Integrated Voltage Measurement... 09:20:25 Magnet Current for MAIN, Imag = 126.3302 A Vtrans = 1.263256 V, Factor V/A = 0.010000 V/A, Zero Current = -0.004689 A 09:20:26 Magnet Current for TRIM, Imag = 6.0032 A Vtrans = 0.500361 V, Factor V/A = 0.083333 V/A, Zero Current = 0.001116 A 09:20:35 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0090000 m VT = 0.060703469 Vs 09:20:43 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0090000 m VT = -0.060725396 Vs 09:20:43 Magnet Current for MAIN, Imag = 126.3305 A Vtrans = 1.263258 V, Factor V/A = 0.010000 V/A, Zero Current = -0.004689 A 09:20:44 Magnet Current for TRIM, Imag = 6.0032 A Vtrans = 0.500360 V, Factor V/A = 0.083333 V/A, Zero Current = 0.001116 A 09:20:52 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0090000 m VT = 0.060700515 Vs 09:21:00 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0090000 m VT = -0.060734311 Vs 09:21:01 Magnet Current for MAIN, Imag = 126.3302 A Vtrans = 1.263256 V, Factor V/A = 0.010000 V/A, Zero Current = -0.004689 A 09:21:01 Magnet Current for TRIM, Imag = 6.0032 A Vtrans = 0.500360 V, Factor V/A = 0.083333 V/A, Zero Current = 0.001116 A 09:21:09 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0090000 m VT = 0.060696452 Vs 09:21:17 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0090000 m VT = -0.060732962 Vs 09:21:18 Magnet Current for MAIN, Imag = 126.3299 A Vtrans = 1.263252 V, Factor V/A = 0.010000 V/A, Zero Current = -0.004689 A 09:21:19 Magnet Current for TRIM, Imag = 6.0032 A Vtrans = 0.500359 V, Factor V/A = 0.083333 V/A, Zero Current = 0.001116 A 09:21:27 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0090000 m VT = 0.060691310 Vs 09:21:34 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0090000 m VT = -0.060737854 Vs 09:21:35 Integrated Voltage Average and RMS of Measurements X0 = 0.0000000 m, Delta X = 0.0090000 m VT = 0.060715284 +- 0.000001233 Vs 09:21:35 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0000000 m, dx = 0.0090000 m, Nturns = 10.0 VT = 0.060715284 +- 0.000001233 Vs BL = 0.67461426 +- 0.00001370 Tm 09:21:38 Position moved: 0.000000 m 09:21:39 BL vs X Fit Coefficients: Cn sigCn n (Tm/m^n) ----- --------+-------- 0 0.67461 0.00001 Compute Fit Residuals: X0 BL sigBL BLfit BL-BLfit (m) (Tm) (Tm) (Tm) (Tm) ----------- --------+-------- -------- -------- 0.0000000 0.67461 0.00001 0.67461 0.00000 09:21:39 Magnet Strength Calculation Harmonic Number = 1 Number Of X0 Positions = 1 X0 BL sigBL (m) (Tm) (Tm) -------- --------+-------- 0.0000 0.67461 0.00001 SL = 0.6746143 +- 0.0000137 Tm^2/m^n 09:21:39 Magnet Strength Measurement Harmonic Number = 1 SL = 0.6746143 +- 0.0000137 Tm^2/m^n 09:21:41 Position moved: 0.000000 m 09:21:42 Magnet system MAIN current ramping to 126.4349 A... 09:21:52 Magnet system TRIM current ramping to 0.0000 A... 09:22:09 Magnet Current for MAIN, Imag = 126.3305 A Vtrans = 1.263258 V, Factor V/A = 0.010000 V/A, Zero Current = -0.004689 A 09:22:09 Magnet Current for TRIM, Imag = 0.0002 A Vtrans = 0.000112 V, Factor V/A = 0.083333 V/A, Zero Current = 0.001116 A 09:22:10 MAIN (V) Voltage Measurement, V = 8.988743 V 09:22:11 Temperature Measurement, Probe Number = 0, T = 27.548000 C 09:22:11 Temperature Measurement, Probe Number = 1, T = 25.884000 C 09:22:12 Temperature Measurement, Probe Number = 2, T = 19.812000 C 09:22:13 Sensor Voltage Measurement, Sensor Number = 0, V = 2.320627 V 09:22:13 Sensor Measurement, Sensor Number = 0, Sensor Type = WATER_HPLEIGHTON_T Raw Sensor Output = 2.320627 (V, A, ohm, ...), Measured Value = 28.342038 (C, atm, Hz, ...) 09:22:13 Sensor Voltage Measurement, Sensor Number = 1, V = 2.306833 V 09:22:13 Sensor Measurement, Sensor Number = 1, Sensor Type = WATER_HPLEIGHTON_T Raw Sensor Output = 2.306833 (V, A, ohm, ...), Measured Value = 28.150457 (C, atm, Hz, ...) 09:22:14 Sensor Voltage Measurement, Sensor Number = 2, V = 2.654752 V 09:22:14 Sensor Measurement, Sensor Number = 2, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 2.654752 (V, A, ohm, ...), Measured Value = 124.106385 (C, atm, Hz, ...) 09:22:15 Sensor Voltage Measurement, Sensor Number = 3, V = 1.093866 V 09:22:15 Sensor Measurement, Sensor Number = 3, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 1.093866 (V, A, ohm, ...), Measured Value = 7.039920 (C, atm, Hz, ...) 09:22:15 Magnet Test Parameter Measurements P0 = MAIN, Value = 126.33 P1 = TRIM, Value = 0.00 P2 = MAIN (V), Value = 8.99 P3 = Coil, Value = 27.55 P4 = Magnet Core, Value = 25.88 P5 = Ambient Temp (C), Value = 19.81 P6 = Supply Temp(C), Value = 28.34 P7 = Return Temp(C), Value = 28.15 P8 = Supply Pres(psi), Value = 124.11 P9 = Return Pres(psi), Value = 7.04 09:22:15 Magnetic Field Measurement Measuring the magnet strength... 09:22:15 Integrated Voltage Measurement... 09:22:15 Magnet Current for MAIN, Imag = 126.3302 A Vtrans = 1.263256 V, Factor V/A = 0.010000 V/A, Zero Current = -0.004689 A 09:22:16 Magnet Current for TRIM, Imag = 0.0002 A Vtrans = 0.000111 V, Factor V/A = 0.083333 V/A, Zero Current = 0.001116 A 09:22:25 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0090000 m VT = 0.058806414 Vs 09:22:33 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0090000 m VT = -0.058858949 Vs 09:22:34 Magnet Current for MAIN, Imag = 126.3299 A Vtrans = 1.263252 V, Factor V/A = 0.010000 V/A, Zero Current = -0.004689 A 09:22:34 Magnet Current for TRIM, Imag = 0.0002 A Vtrans = 0.000110 V, Factor V/A = 0.083333 V/A, Zero Current = 0.001116 A 09:22:42 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0090000 m VT = 0.058833861 Vs 09:22:50 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0090000 m VT = -0.058858278 Vs 09:22:51 Magnet Current for MAIN, Imag = 126.3299 A Vtrans = 1.263252 V, Factor V/A = 0.010000 V/A, Zero Current = -0.004689 A 09:22:51 Magnet Current for TRIM, Imag = 0.0003 A Vtrans = 0.000114 V, Factor V/A = 0.083333 V/A, Zero Current = 0.001116 A 09:22:59 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0090000 m VT = 0.058835836 Vs 09:23:07 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0090000 m VT = -0.058855577 Vs 09:23:08 Magnet Current for MAIN, Imag = 126.3301 A Vtrans = 1.263254 V, Factor V/A = 0.010000 V/A, Zero Current = -0.004689 A 09:23:09 Magnet Current for TRIM, Imag = 0.0002 A Vtrans = 0.000111 V, Factor V/A = 0.083333 V/A, Zero Current = 0.001116 A 09:23:17 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0090000 m VT = 0.058826536 Vs 09:23:25 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0090000 m VT = -0.058847112 Vs 09:23:25 Integrated Voltage Average and RMS of Measurements X0 = 0.0000000 m, Delta X = 0.0090000 m VT = 0.058840321 +- 0.000005759 Vs 09:23:25 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0000000 m, dx = 0.0090000 m, Nturns = 10.0 VT = 0.058840321 +- 0.000005759 Vs BL = 0.65378134 +- 0.00006398 Tm 09:23:29 Position moved: 0.000000 m 09:23:29 BL vs X Fit Coefficients: Cn sigCn n (Tm/m^n) ----- --------+-------- 0 0.65378 0.00006 Compute Fit Residuals: X0 BL sigBL BLfit BL-BLfit (m) (Tm) (Tm) (Tm) (Tm) ----------- --------+-------- -------- -------- 0.0000000 0.65378 0.00006 0.65378 0.00000 09:23:29 Magnet Strength Calculation Harmonic Number = 1 Number Of X0 Positions = 1 X0 BL sigBL (m) (Tm) (Tm) -------- --------+-------- 0.0000 0.65378 0.00006 SL = 0.6537813 +- 0.0000640 Tm^2/m^n 09:23:29 Magnet Strength Measurement Harmonic Number = 1 SL = 0.6537813 +- 0.0000640 Tm^2/m^n 09:23:32 Position moved: 0.000000 m 09:23:32 Magnet system MAIN current ramping to 126.4349 A... 09:23:42 Magnet system TRIM current ramping to -6.0000 A... 09:23:59 Magnet Current for MAIN, Imag = 126.3299 A Vtrans = 1.263252 V, Factor V/A = 0.010000 V/A, Zero Current = -0.004689 A 09:24:00 Magnet Current for TRIM, Imag = -6.0039 A Vtrans = -0.500232 V, Factor V/A = 0.083333 V/A, Zero Current = 0.001116 A 09:24:00 MAIN (V) Voltage Measurement, V = 8.988965 V 09:24:01 Temperature Measurement, Probe Number = 0, T = 27.620000 C 09:24:01 Temperature Measurement, Probe Number = 1, T = 25.950000 C 09:24:02 Temperature Measurement, Probe Number = 2, T = 19.848000 C 09:24:03 Sensor Voltage Measurement, Sensor Number = 0, V = 2.321645 V 09:24:03 Sensor Measurement, Sensor Number = 0, Sensor Type = WATER_HPLEIGHTON_T Raw Sensor Output = 2.321645 (V, A, ohm, ...), Measured Value = 28.356178 (C, atm, Hz, ...) 09:24:03 Sensor Voltage Measurement, Sensor Number = 1, V = 2.310316 V 09:24:03 Sensor Measurement, Sensor Number = 1, Sensor Type = WATER_HPLEIGHTON_T Raw Sensor Output = 2.310316 (V, A, ohm, ...), Measured Value = 28.198829 (C, atm, Hz, ...) 09:24:04 Sensor Voltage Measurement, Sensor Number = 2, V = 2.663565 V 09:24:04 Sensor Measurement, Sensor Number = 2, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 2.663565 (V, A, ohm, ...), Measured Value = 124.767405 (C, atm, Hz, ...) 09:24:05 Sensor Voltage Measurement, Sensor Number = 3, V = 1.093863 V 09:24:05 Sensor Measurement, Sensor Number = 3, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 1.093863 (V, A, ohm, ...), Measured Value = 7.039725 (C, atm, Hz, ...) 09:24:05 Magnet Test Parameter Measurements P0 = MAIN, Value = 126.33 P1 = TRIM, Value = -6.00 P2 = MAIN (V), Value = 8.99 P3 = Coil, Value = 27.62 P4 = Magnet Core, Value = 25.95 P5 = Ambient Temp (C), Value = 19.85 P6 = Supply Temp(C), Value = 28.36 P7 = Return Temp(C), Value = 28.20 P8 = Supply Pres(psi), Value = 124.77 P9 = Return Pres(psi), Value = 7.04 09:24:05 Magnetic Field Measurement Measuring the magnet strength... 09:24:05 Integrated Voltage Measurement... 09:24:06 Magnet Current for MAIN, Imag = 126.3301 A Vtrans = 1.263254 V, Factor V/A = 0.010000 V/A, Zero Current = -0.004689 A 09:24:06 Magnet Current for TRIM, Imag = -6.0039 A Vtrans = -0.500232 V, Factor V/A = 0.083333 V/A, Zero Current = 0.001116 A 09:24:15 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0090000 m VT = 0.056924998 Vs 09:24:23 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0090000 m VT = -0.056961558 Vs 09:24:24 Magnet Current for MAIN, Imag = 126.3301 A Vtrans = 1.263254 V, Factor V/A = 0.010000 V/A, Zero Current = -0.004689 A 09:24:24 Magnet Current for TRIM, Imag = -6.0039 A Vtrans = -0.500231 V, Factor V/A = 0.083333 V/A, Zero Current = 0.001116 A 09:24:32 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0090000 m VT = 0.056929396 Vs 09:24:40 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0090000 m VT = -0.056975001 Vs 09:24:41 Magnet Current for MAIN, Imag = 126.3300 A Vtrans = 1.263253 V, Factor V/A = 0.010000 V/A, Zero Current = -0.004689 A 09:24:41 Magnet Current for TRIM, Imag = -6.0039 A Vtrans = -0.500229 V, Factor V/A = 0.083333 V/A, Zero Current = 0.001116 A 09:24:49 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0090000 m VT = 0.056919477 Vs 09:24:57 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0090000 m VT = -0.056970757 Vs 09:24:58 Magnet Current for MAIN, Imag = 126.3299 A Vtrans = 1.263252 V, Factor V/A = 0.010000 V/A, Zero Current = -0.004689 A 09:24:59 Magnet Current for TRIM, Imag = -6.0039 A Vtrans = -0.500229 V, Factor V/A = 0.083333 V/A, Zero Current = 0.001116 A 09:25:06 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0090000 m VT = 0.056919011 Vs 09:25:14 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0090000 m VT = -0.056969811 Vs 09:25:14 Integrated Voltage Average and RMS of Measurements X0 = 0.0000000 m, Delta X = 0.0090000 m VT = 0.056946251 +- 0.000003496 Vs 09:25:14 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0000000 m, dx = 0.0090000 m, Nturns = 10.0 VT = 0.056946251 +- 0.000003496 Vs BL = 0.63273612 +- 0.00003885 Tm 09:25:18 Position moved: 0.000000 m 09:25:18 BL vs X Fit Coefficients: Cn sigCn n (Tm/m^n) ----- --------+-------- 0 0.63274 0.00004 Compute Fit Residuals: X0 BL sigBL BLfit BL-BLfit (m) (Tm) (Tm) (Tm) (Tm) ----------- --------+-------- -------- -------- 0.0000000 0.63274 0.00004 0.63274 0.00000 09:25:18 Magnet Strength Calculation Harmonic Number = 1 Number Of X0 Positions = 1 X0 BL sigBL (m) (Tm) (Tm) -------- --------+-------- 0.0000 0.63274 0.00004 SL = 0.6327361 +- 0.0000388 Tm^2/m^n 09:25:18 Magnet Strength Measurement Harmonic Number = 1 SL = 0.6327361 +- 0.0000388 Tm^2/m^n 09:25:21 Position moved: 0.000000 m 09:25:21 Magnet system MAIN current ramping to 126.4349 A... 09:25:32 Magnet system TRIM current ramping to 6.0000 A... 09:25:55 Magnet Current for MAIN, Imag = 126.3299 A Vtrans = 1.263252 V, Factor V/A = 0.010000 V/A, Zero Current = -0.004689 A 09:25:55 Magnet Current for TRIM, Imag = 6.0032 A Vtrans = 0.500360 V, Factor V/A = 0.083333 V/A, Zero Current = 0.001116 A 09:25:56 MAIN (V) Voltage Measurement, V = 8.989261 V 09:25:57 Temperature Measurement, Probe Number = 0, T = 27.627000 C 09:25:57 Temperature Measurement, Probe Number = 1, T = 25.918000 C 09:25:58 Temperature Measurement, Probe Number = 2, T = 19.802000 C 09:25:58 Sensor Voltage Measurement, Sensor Number = 0, V = 2.320726 V 09:25:58 Sensor Measurement, Sensor Number = 0, Sensor Type = WATER_HPLEIGHTON_T Raw Sensor Output = 2.320726 (V, A, ohm, ...), Measured Value = 28.343414 (C, atm, Hz, ...) 09:25:59 Sensor Voltage Measurement, Sensor Number = 1, V = 2.312880 V 09:25:59 Sensor Measurement, Sensor Number = 1, Sensor Type = WATER_HPLEIGHTON_T Raw Sensor Output = 2.312880 (V, A, ohm, ...), Measured Value = 28.234438 (C, atm, Hz, ...) 09:26:00 Sensor Voltage Measurement, Sensor Number = 2, V = 2.662596 V 09:26:00 Sensor Measurement, Sensor Number = 2, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 2.662596 (V, A, ohm, ...), Measured Value = 124.694670 (C, atm, Hz, ...) 09:26:00 Sensor Voltage Measurement, Sensor Number = 3, V = 1.093966 V 09:26:00 Sensor Measurement, Sensor Number = 3, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 1.093966 (V, A, ohm, ...), Measured Value = 7.047450 (C, atm, Hz, ...) 09:26:01 Magnet Test Parameter Measurements P0 = MAIN, Value = 126.33 P1 = TRIM, Value = 6.00 P2 = MAIN (V), Value = 8.99 P3 = Coil, Value = 27.63 P4 = Magnet Core, Value = 25.92 P5 = Ambient Temp (C), Value = 19.80 P6 = Supply Temp(C), Value = 28.34 P7 = Return Temp(C), Value = 28.23 P8 = Supply Pres(psi), Value = 124.69 P9 = Return Pres(psi), Value = 7.05 09:26:01 Magnetic Field Measurement Measuring the magnet strength... 09:26:01 Integrated Voltage Measurement... 09:26:01 Magnet Current for MAIN, Imag = 126.3299 A Vtrans = 1.263252 V, Factor V/A = 0.010000 V/A, Zero Current = -0.004689 A 09:26:02 Magnet Current for TRIM, Imag = 6.0032 A Vtrans = 0.500360 V, Factor V/A = 0.083333 V/A, Zero Current = 0.001116 A 09:26:11 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0090000 m VT = 0.060673511 Vs 09:26:19 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0090000 m VT = -0.060735780 Vs 09:26:19 Magnet Current for MAIN, Imag = 126.3299 A Vtrans = 1.263252 V, Factor V/A = 0.010000 V/A, Zero Current = -0.004689 A 09:26:20 Magnet Current for TRIM, Imag = 6.0032 A Vtrans = 0.500358 V, Factor V/A = 0.083333 V/A, Zero Current = 0.001116 A 09:26:28 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0090000 m VT = 0.060693218 Vs 09:26:36 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0090000 m VT = -0.060728942 Vs 09:26:36 Magnet Current for MAIN, Imag = 126.3296 A Vtrans = 1.263249 V, Factor V/A = 0.010000 V/A, Zero Current = -0.004689 A 09:26:37 Magnet Current for TRIM, Imag = 6.0032 A Vtrans = 0.500357 V, Factor V/A = 0.083333 V/A, Zero Current = 0.001116 A 09:26:45 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0090000 m VT = 0.060694556 Vs 09:26:53 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0090000 m VT = -0.060731468 Vs 09:26:54 Magnet Current for MAIN, Imag = 126.3295 A Vtrans = 1.263248 V, Factor V/A = 0.010000 V/A, Zero Current = -0.004689 A 09:26:54 Magnet Current for TRIM, Imag = 6.0032 A Vtrans = 0.500356 V, Factor V/A = 0.083333 V/A, Zero Current = 0.001116 A 09:27:02 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0090000 m VT = 0.060695739 Vs 09:27:10 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0090000 m VT = -0.060729242 Vs 09:27:10 Integrated Voltage Average and RMS of Measurements X0 = 0.0000000 m, Delta X = 0.0090000 m VT = 0.060710307 +- 0.000003344 Vs 09:27:10 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0000000 m, dx = 0.0090000 m, Nturns = 10.0 VT = 0.060710307 +- 0.000003344 Vs BL = 0.67455897 +- 0.00003716 Tm 09:27:14 Position moved: 0.000000 m 09:27:14 BL vs X Fit Coefficients: Cn sigCn n (Tm/m^n) ----- --------+-------- 0 0.67456 0.00004 Compute Fit Residuals: X0 BL sigBL BLfit BL-BLfit (m) (Tm) (Tm) (Tm) (Tm) ----------- --------+-------- -------- -------- 0.0000000 0.67456 0.00004 0.67456 0.00000 09:27:14 Magnet Strength Calculation Harmonic Number = 1 Number Of X0 Positions = 1 X0 BL sigBL (m) (Tm) (Tm) -------- --------+-------- 0.0000 0.67456 0.00004 SL = 0.6745590 +- 0.0000372 Tm^2/m^n 09:27:14 Magnet Strength Measurement Harmonic Number = 1 SL = 0.6745590 +- 0.0000372 Tm^2/m^n 09:27:17 Position moved: 0.000000 m 09:27:20 Position moved: 0.000000 m 09:27:23 Magnet system MAIN current ramping to 0.0000 A... 09:28:34 Magnet system TRIM current ramping to 0.0000 A...