SLAC Magnetic Measurements Date: 10-04-2022 Time: 09:03:58 Project: LCLS-II-HE Device Type: Undulator Data Type: Serial Number: HE_SXU_000 DATASET Number: 1 Tuning Stage: Tuning Run Number: 75 Measurement Type: Long Coil Scans Operator: YL Comment: Skew Quad, 4H Center position: X0=0.000000, Y0=0.000000 Nominal Gap=7.20 Scan Raw value Integral Nominal Actual Pre-measurement Gap Actual Post-measurement Gap Number Volt*sec T*m | T*m^2 Gap US - mm US Raw DS - mm DS Raw US - mm US Raw DS - mm DS Raw --------------------------------------------------------------------------------------------------------------------------------------- 0 -3.916E-06 -55.936E-06 7.20000 7.19765 0 7.19905 0 7.19765 0 7.19900 0 1 -4.036E-06 -57.648E-06 7.20000 7.19760 0 7.19895 0 7.19755 0 7.19895 0 2 -4.098E-06 -58.534E-06 7.20000 7.19750 0 7.19890 0 7.19745 0 7.19890 0 3 -4.043E-06 -57.741E-06 7.20000 7.19740 0 7.19885 0 7.19740 0 7.19885 0 4 -4.086E-06 -58.360E-06 7.20000 7.19740 0 7.19880 0 7.19735 0 7.19880 0