SLAC Magnetic Measurements Date: 01-17-2025 Time: 14:48:35 Project: LBL Magnet Type: Quad Magnet Name: TQFC_Q11_BTA_QF8 Serial Number: AL-1423-8026-0001 Run Number: 2 Measurement Device: 10wire Operator: sda Comment: Integrated strength of By Trim at +/-25 Test Currents for HTRIM (A): 25.00 25.00 25.00 -25.00 -25.00 -25.00 14:49:03 Position moved: 0.000000 m 14:49:08 Magnet Current for HTRIM, Imag = 0.0013 A Vtrans = 0.000111 V, Factor V/A = 0.083333 V/A, Zero Current = 0.000000 A 14:49:09 Magnet Current for HTRIM, Imag = 0.0013 A Vtrans = 0.000111 V, Factor V/A = 0.083333 V/A, Zero Current = 0.000000 A 14:49:10 Magnet Current for HTRIM, Imag = 0.0013 A Vtrans = 0.000111 V, Factor V/A = 0.083333 V/A, Zero Current = 0.000000 A 14:49:11 Magnet Current for HTRIM, Imag = 0.0013 A Vtrans = 0.000111 V, Factor V/A = 0.083333 V/A, Zero Current = 0.000000 A 14:49:11 Magnet Avgerage Current Measurement for HTRIM : Imag = 0.0013 +- 0.0000 A 14:49:27 Initial Zero Current Measurement for HTRIM, Imag_zero = 0.00133295 A 14:49:27 Ambient Measurement Device(s) Being Used, type = None 14:49:27 Magnet Voltage Measurement Device(s) Being Used, type = HP34970 14:49:28 Temperature Measurement Device(s) Being Used, type = HPTHERM_HP34970 14:49:28 XMAG Device(s) Being Used, type = None 14:49:28 Ambient Measurement Device(s) Being Used, type = HP34970 14:49:31 Position moved: 0.000000 m 14:49:34 Magnet system HTRIM current ramping to 25.0000 A... 14:50:05 Magnet Current for HTRIM, Imag = 24.9870 A Vtrans = 2.082363 V, Factor V/A = 0.083333 V/A, Zero Current = 0.001333 A 14:50:06 Main (V) Voltage Measurement, V = 0.000001 V 14:50:13 Temperature Measurement, Probe Number = 0, T = 21.910000 C 14:50:13 Temperature Measurement, Probe Number = 1, T = 24.006000 C 14:50:14 Temperature Measurement, Probe Number = 2, T = 28.114000 C 14:50:15 Temperature Measurement, Probe Number = 3, T = 27.017000 C 14:50:15 Temperature Measurement, Probe Number = 4, T = 17.367000 C 14:50:16 Sensor Voltage Measurement, Sensor Number = 0, V = 2.644909 V 14:50:16 Sensor Measurement, Sensor Number = 0, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 2.644909 (V, A, ohm, ...), Measured Value = 123.368212 (C, atm, Hz, ...) 14:50:17 Sensor Voltage Measurement, Sensor Number = 1, V = 1.171712 V 14:50:17 Sensor Measurement, Sensor Number = 1, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 1.171712 (V, A, ohm, ...), Measured Value = 12.878430 (C, atm, Hz, ...) 14:50:18 Sensor Voltage Measurement, Sensor Number = 2, V = 0.001534 V 14:50:18 Sensor Measurement, Sensor Number = 2, Sensor Type = WATER_KEYENCE_FLOW Raw Sensor Output = 0.001534 (V, A, ohm, ...), Measured Value = 1.533593 (C, atm, Hz, ...) 14:50:18 Magnet Test Parameter Measurements P0 = HTRIM, Value = 24.99 P1 = Main (V), Value = 0.00 P2 = Core_Top, Value = 21.91 P3 = Top_Coil, Value = 24.01 P4 = H20 In, Value = 28.11 P5 = H20 Out, Value = 27.02 P6 = Ambient, Value = 17.37 P7 = Supply Pres(psi), Value = 123.37 P8 = Return Pres(psi), Value = 12.88 P9 = Flow Rate(gpm), Value = 1.53 14:50:18 Magnetic Field Measurement Measuring the magnet strength... 14:50:18 Integrated Voltage Measurement... 14:50:19 Magnet Current for HTRIM, Imag = 24.9875 A Vtrans = 2.082399 V, Factor V/A = 0.083333 V/A, Zero Current = 0.001333 A 14:50:27 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0200000 m VT = 0.001344397 Vs 14:50:34 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0200000 m VT = -0.001341024 Vs 14:50:35 Magnet Current for HTRIM, Imag = 24.9881 A Vtrans = 2.082456 V, Factor V/A = 0.083333 V/A, Zero Current = 0.001333 A 14:50:43 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0200000 m VT = 0.001342203 Vs 14:50:50 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0200000 m VT = -0.001343310 Vs 14:50:51 Magnet Current for HTRIM, Imag = 24.9884 A Vtrans = 2.082481 V, Factor V/A = 0.083333 V/A, Zero Current = 0.001333 A 14:50:58 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0200000 m VT = 0.001343064 Vs 14:51:06 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0200000 m VT = -0.001342127 Vs 14:51:06 Magnet Current for HTRIM, Imag = 24.9886 A Vtrans = 2.082494 V, Factor V/A = 0.083333 V/A, Zero Current = 0.001333 A 14:51:14 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0200000 m VT = 0.001343798 Vs 14:51:21 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0200000 m VT = -0.001342333 Vs 14:51:22 Magnet Current for HTRIM, Imag = 24.9888 A Vtrans = 2.082515 V, Factor V/A = 0.083333 V/A, Zero Current = 0.001333 A 14:51:29 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0200000 m VT = 0.001343386 Vs 14:51:37 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0200000 m VT = -0.001342993 Vs 14:51:37 Integrated Voltage Average and RMS of Measurements X0 = 0.0000000 m, Delta X = 0.0200000 m VT = 0.001342863 +- 0.000000225 Vs 14:51:37 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0000000 m, dx = 0.0200000 m, Nturns = 10.0 VT = 0.001342863 +- 0.000000225 Vs BL = 0.00671432 +- 0.00000113 Tm 14:51:41 Position moved: 0.000000 m 14:51:41 BL vs X Fit Coefficients: Cn sigCn n (Tm/m^n) ----- --------+-------- 0 0.00671 0.00000 Compute Fit Residuals: X0 BL sigBL BLfit BL-BLfit (m) (Tm) (Tm) (Tm) (Tm) ----------- --------+-------- -------- -------- 0.0000000 0.00671 0.00000 0.00671 0.00000 14:51:41 Magnet Strength Calculation Harmonic Number = 1 Number Of X0 Positions = 1 X0 BL sigBL (m) (Tm) (Tm) -------- --------+-------- 0.0000 0.00671 0.00000 SL = 0.0067143 +- 0.0000011 Tm^2/m^n 14:51:41 Magnet Strength Measurement Harmonic Number = 1 SL = 0.0067143 +- 0.0000011 Tm^2/m^n 14:51:44 Position moved: 0.000000 m 14:51:45 Magnet Current for HTRIM, Imag = 24.9890 A Vtrans = 2.082526 V, Factor V/A = 0.083333 V/A, Zero Current = 0.001333 A 14:51:46 Main (V) Voltage Measurement, V = 0.000002 V 14:51:49 Temperature Measurement, Probe Number = 0, T = 21.993000 C 14:51:50 Temperature Measurement, Probe Number = 1, T = 24.063000 C 14:51:51 Temperature Measurement, Probe Number = 2, T = 28.107000 C 14:51:51 Temperature Measurement, Probe Number = 3, T = 26.978000 C 14:51:52 Temperature Measurement, Probe Number = 4, T = 17.399000 C 14:51:53 Sensor Voltage Measurement, Sensor Number = 0, V = 2.649776 V 14:51:53 Sensor Measurement, Sensor Number = 0, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 2.649776 (V, A, ohm, ...), Measured Value = 123.733163 (C, atm, Hz, ...) 14:51:53 Sensor Voltage Measurement, Sensor Number = 1, V = 1.171866 V 14:51:53 Sensor Measurement, Sensor Number = 1, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 1.171866 (V, A, ohm, ...), Measured Value = 12.889950 (C, atm, Hz, ...) 14:51:54 Sensor Voltage Measurement, Sensor Number = 2, V = 0.001503 V 14:51:54 Sensor Measurement, Sensor Number = 2, Sensor Type = WATER_KEYENCE_FLOW Raw Sensor Output = 0.001503 (V, A, ohm, ...), Measured Value = 1.503190 (C, atm, Hz, ...) 14:51:54 Magnet Test Parameter Measurements P0 = HTRIM, Value = 24.99 P1 = Main (V), Value = 0.00 P2 = Core_Top, Value = 21.99 P3 = Top_Coil, Value = 24.06 P4 = H20 In, Value = 28.11 P5 = H20 Out, Value = 26.98 P6 = Ambient, Value = 17.40 P7 = Supply Pres(psi), Value = 123.73 P8 = Return Pres(psi), Value = 12.89 P9 = Flow Rate(gpm), Value = 1.50 14:51:54 Magnet system HTRIM current ramping to 25.0000 A... 14:52:16 Magnet Current for HTRIM, Imag = 24.9893 A Vtrans = 2.082554 V, Factor V/A = 0.083333 V/A, Zero Current = 0.001333 A 14:52:16 Main (V) Voltage Measurement, V = 0.000002 V 14:52:20 Temperature Measurement, Probe Number = 0, T = 22.132000 C 14:52:21 Temperature Measurement, Probe Number = 1, T = 24.120000 C 14:52:21 Temperature Measurement, Probe Number = 2, T = 28.108000 C 14:52:22 Temperature Measurement, Probe Number = 3, T = 26.970000 C 14:52:23 Temperature Measurement, Probe Number = 4, T = 17.383000 C 14:52:23 Sensor Voltage Measurement, Sensor Number = 0, V = 2.642212 V 14:52:23 Sensor Measurement, Sensor Number = 0, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 2.642212 (V, A, ohm, ...), Measured Value = 123.165930 (C, atm, Hz, ...) 14:52:24 Sensor Voltage Measurement, Sensor Number = 1, V = 1.170823 V 14:52:24 Sensor Measurement, Sensor Number = 1, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 1.170823 (V, A, ohm, ...), Measured Value = 12.811747 (C, atm, Hz, ...) 14:52:25 Sensor Voltage Measurement, Sensor Number = 2, V = 0.001523 V 14:52:25 Sensor Measurement, Sensor Number = 2, Sensor Type = WATER_KEYENCE_FLOW Raw Sensor Output = 0.001523 (V, A, ohm, ...), Measured Value = 1.522846 (C, atm, Hz, ...) 14:52:25 Magnet Test Parameter Measurements P0 = HTRIM, Value = 24.99 P1 = Main (V), Value = 0.00 P2 = Core_Top, Value = 22.13 P3 = Top_Coil, Value = 24.12 P4 = H20 In, Value = 28.11 P5 = H20 Out, Value = 26.97 P6 = Ambient, Value = 17.38 P7 = Supply Pres(psi), Value = 123.17 P8 = Return Pres(psi), Value = 12.81 P9 = Flow Rate(gpm), Value = 1.52 14:52:25 Magnetic Field Measurement Measuring the magnet strength... 14:52:25 Integrated Voltage Measurement... 14:52:26 Magnet Current for HTRIM, Imag = 24.9892 A Vtrans = 2.082541 V, Factor V/A = 0.083333 V/A, Zero Current = 0.001333 A 14:52:34 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0200000 m VT = 0.001340041 Vs 14:52:41 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0200000 m VT = -0.001344186 Vs 14:52:42 Magnet Current for HTRIM, Imag = 24.9894 A Vtrans = 2.082559 V, Factor V/A = 0.083333 V/A, Zero Current = 0.001333 A 14:52:50 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0200000 m VT = 0.001342065 Vs 14:52:57 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0200000 m VT = -0.001343882 Vs 14:52:58 Magnet Current for HTRIM, Imag = 24.9900 A Vtrans = 2.082614 V, Factor V/A = 0.083333 V/A, Zero Current = 0.001333 A 14:53:05 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0200000 m VT = 0.001341571 Vs 14:53:12 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0200000 m VT = -0.001343139 Vs 14:53:13 Magnet Current for HTRIM, Imag = 24.9902 A Vtrans = 2.082625 V, Factor V/A = 0.083333 V/A, Zero Current = 0.001333 A 14:53:21 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0200000 m VT = 0.001340669 Vs 14:53:28 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0200000 m VT = -0.001345116 Vs 14:53:29 Magnet Current for HTRIM, Imag = 24.9904 A Vtrans = 2.082644 V, Factor V/A = 0.083333 V/A, Zero Current = 0.001333 A 14:53:37 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0200000 m VT = 0.001339766 Vs 14:53:44 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0200000 m VT = -0.001344837 Vs 14:53:44 Integrated Voltage Average and RMS of Measurements X0 = 0.0000000 m, Delta X = 0.0200000 m VT = 0.001342527 +- 0.000000342 Vs 14:53:44 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0000000 m, dx = 0.0200000 m, Nturns = 10.0 VT = 0.001342527 +- 0.000000342 Vs BL = 0.00671264 +- 0.00000171 Tm 14:53:48 Position moved: 0.000000 m 14:53:48 BL vs X Fit Coefficients: Cn sigCn n (Tm/m^n) ----- --------+-------- 0 0.00671 0.00000 Compute Fit Residuals: X0 BL sigBL BLfit BL-BLfit (m) (Tm) (Tm) (Tm) (Tm) ----------- --------+-------- -------- -------- 0.0000000 0.00671 0.00000 0.00671 0.00000 14:53:48 Magnet Strength Calculation Harmonic Number = 1 Number Of X0 Positions = 1 X0 BL sigBL (m) (Tm) (Tm) -------- --------+-------- 0.0000 0.00671 0.00000 SL = 0.0067126 +- 0.0000017 Tm^2/m^n 14:53:48 Magnet Strength Measurement Harmonic Number = 1 SL = 0.0067126 +- 0.0000017 Tm^2/m^n Remeasure 14:53:48 Magnetic Field Measurement Measuring the magnet strength... 14:53:48 Integrated Voltage Measurement... 14:53:49 Magnet Current for HTRIM, Imag = 24.9906 A Vtrans = 2.082665 V, Factor V/A = 0.083333 V/A, Zero Current = 0.001333 A 14:53:58 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0200000 m VT = 0.001340707 Vs 14:54:05 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0200000 m VT = -0.001344568 Vs 14:54:06 Magnet Current for HTRIM, Imag = 24.9908 A Vtrans = 2.082679 V, Factor V/A = 0.083333 V/A, Zero Current = 0.001333 A 14:54:13 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0200000 m VT = 0.001340398 Vs 14:54:20 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0200000 m VT = -0.001343959 Vs 14:54:21 Magnet Current for HTRIM, Imag = 24.9909 A Vtrans = 2.082682 V, Factor V/A = 0.083333 V/A, Zero Current = 0.001333 A 14:54:29 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0200000 m VT = 0.001341232 Vs 14:54:36 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0200000 m VT = -0.001345769 Vs 14:54:37 Magnet Current for HTRIM, Imag = 24.9909 A Vtrans = 2.082689 V, Factor V/A = 0.083333 V/A, Zero Current = 0.001333 A 14:54:44 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0200000 m VT = 0.001340301 Vs 14:54:52 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0200000 m VT = -0.001345349 Vs 14:54:52 Magnet Current for HTRIM, Imag = 24.9911 A Vtrans = 2.082702 V, Factor V/A = 0.083333 V/A, Zero Current = 0.001333 A 14:55:00 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0200000 m VT = 0.001339971 Vs 14:55:07 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0200000 m VT = -0.001343491 Vs 14:55:07 Integrated Voltage Average and RMS of Measurements X0 = 0.0000000 m, Delta X = 0.0200000 m VT = 0.001342574 +- 0.000000599 Vs 14:55:07 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0000000 m, dx = 0.0200000 m, Nturns = 10.0 VT = 0.001342574 +- 0.000000599 Vs BL = 0.00671287 +- 0.00000299 Tm 14:55:11 Position moved: 0.000000 m 14:55:11 BL vs X Fit Coefficients: Cn sigCn n (Tm/m^n) ----- --------+-------- 0 0.00671 0.00000 Compute Fit Residuals: X0 BL sigBL BLfit BL-BLfit (m) (Tm) (Tm) (Tm) (Tm) ----------- --------+-------- -------- -------- 0.0000000 0.00671 0.00000 0.00671 0.00000 14:55:11 Magnet Strength Calculation Harmonic Number = 1 Number Of X0 Positions = 1 X0 BL sigBL (m) (Tm) (Tm) -------- --------+-------- 0.0000 0.00671 0.00000 SL = 0.0067129 +- 0.0000030 Tm^2/m^n 14:55:11 Magnet Strength Measurement Harmonic Number = 1 SL = 0.0067129 +- 0.0000030 Tm^2/m^n Remeasure 14:55:12 Magnetic Field Measurement Measuring the magnet strength... 14:55:12 Integrated Voltage Measurement... 14:55:13 Magnet Current for HTRIM, Imag = 24.9912 A Vtrans = 2.082714 V, Factor V/A = 0.083333 V/A, Zero Current = 0.001333 A 14:55:21 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0200000 m VT = 0.001340928 Vs 14:55:29 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0200000 m VT = -0.001344064 Vs 14:55:29 Magnet Current for HTRIM, Imag = 24.9912 A Vtrans = 2.082710 V, Factor V/A = 0.083333 V/A, Zero Current = 0.001333 A 14:55:37 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0200000 m VT = 0.001341514 Vs 14:55:44 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0200000 m VT = -0.001344394 Vs 14:55:45 Magnet Current for HTRIM, Imag = 24.9913 A Vtrans = 2.082719 V, Factor V/A = 0.083333 V/A, Zero Current = 0.001333 A 14:55:53 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0200000 m VT = 0.001341283 Vs 14:56:01 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0200000 m VT = -0.001342907 Vs 14:56:02 Magnet Current for HTRIM, Imag = 24.9912 A Vtrans = 2.082709 V, Factor V/A = 0.083333 V/A, Zero Current = 0.001333 A 14:56:10 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0200000 m VT = 0.001341725 Vs 14:56:17 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0200000 m VT = -0.001343817 Vs 14:56:18 Magnet Current for HTRIM, Imag = 24.9912 A Vtrans = 2.082711 V, Factor V/A = 0.083333 V/A, Zero Current = 0.001333 A 14:56:26 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0200000 m VT = 0.001343132 Vs 14:56:33 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0200000 m VT = -0.001342442 Vs 14:56:33 Integrated Voltage Average and RMS of Measurements X0 = 0.0000000 m, Delta X = 0.0200000 m VT = 0.001342621 +- 0.000000301 Vs 14:56:33 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0000000 m, dx = 0.0200000 m, Nturns = 10.0 VT = 0.001342621 +- 0.000000301 Vs BL = 0.00671310 +- 0.00000151 Tm 14:56:37 Position moved: 0.000000 m 14:56:37 BL vs X Fit Coefficients: Cn sigCn n (Tm/m^n) ----- --------+-------- 0 0.00671 0.00000 Compute Fit Residuals: X0 BL sigBL BLfit BL-BLfit (m) (Tm) (Tm) (Tm) (Tm) ----------- --------+-------- -------- -------- 0.0000000 0.00671 0.00000 0.00671 0.00000 14:56:37 Magnet Strength Calculation Harmonic Number = 1 Number Of X0 Positions = 1 X0 BL sigBL (m) (Tm) (Tm) -------- --------+-------- 0.0000 0.00671 0.00000 SL = 0.0067131 +- 0.0000015 Tm^2/m^n 14:56:37 Magnet Strength Measurement Harmonic Number = 1 SL = 0.0067131 +- 0.0000015 Tm^2/m^n 14:56:41 Position moved: 0.000000 m 14:56:41 Magnet Current for HTRIM, Imag = 24.9912 A Vtrans = 2.082710 V, Factor V/A = 0.083333 V/A, Zero Current = 0.001333 A 14:56:42 Main (V) Voltage Measurement, V = 0.000002 V 14:56:46 Temperature Measurement, Probe Number = 0, T = 22.480000 C 14:56:46 Temperature Measurement, Probe Number = 1, T = 24.608000 C 14:56:47 Temperature Measurement, Probe Number = 2, T = 28.103000 C 14:56:48 Temperature Measurement, Probe Number = 3, T = 26.936000 C 14:56:48 Temperature Measurement, Probe Number = 4, T = 17.420000 C 14:56:49 Sensor Voltage Measurement, Sensor Number = 0, V = 2.665949 V 14:56:49 Sensor Measurement, Sensor Number = 0, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 2.665949 (V, A, ohm, ...), Measured Value = 124.946213 (C, atm, Hz, ...) 14:56:50 Sensor Voltage Measurement, Sensor Number = 1, V = 1.170945 V 14:56:50 Sensor Measurement, Sensor Number = 1, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 1.170945 (V, A, ohm, ...), Measured Value = 12.820838 (C, atm, Hz, ...) 14:56:51 Sensor Voltage Measurement, Sensor Number = 2, V = 0.001503 V 14:56:51 Sensor Measurement, Sensor Number = 2, Sensor Type = WATER_KEYENCE_FLOW Raw Sensor Output = 0.001503 (V, A, ohm, ...), Measured Value = 1.503204 (C, atm, Hz, ...) 14:56:51 Magnet Test Parameter Measurements P0 = HTRIM, Value = 24.99 P1 = Main (V), Value = 0.00 P2 = Core_Top, Value = 22.48 P3 = Top_Coil, Value = 24.61 P4 = H20 In, Value = 28.10 P5 = H20 Out, Value = 26.94 P6 = Ambient, Value = 17.42 P7 = Supply Pres(psi), Value = 124.95 P8 = Return Pres(psi), Value = 12.82 P9 = Flow Rate(gpm), Value = 1.50 14:56:51 Magnet system HTRIM current ramping to 25.0000 A... 14:57:12 Magnet Current for HTRIM, Imag = 24.9911 A Vtrans = 2.082701 V, Factor V/A = 0.083333 V/A, Zero Current = 0.001333 A 14:57:13 Main (V) Voltage Measurement, V = 0.000002 V 14:57:17 Temperature Measurement, Probe Number = 0, T = 22.531000 C 14:57:17 Temperature Measurement, Probe Number = 1, T = 24.653000 C 14:57:18 Temperature Measurement, Probe Number = 2, T = 28.106000 C 14:57:18 Temperature Measurement, Probe Number = 3, T = 26.934000 C 14:57:19 Temperature Measurement, Probe Number = 4, T = 17.419000 C 14:57:20 Sensor Voltage Measurement, Sensor Number = 0, V = 2.639039 V 14:57:20 Sensor Measurement, Sensor Number = 0, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 2.639039 (V, A, ohm, ...), Measured Value = 122.927962 (C, atm, Hz, ...) 14:57:21 Sensor Voltage Measurement, Sensor Number = 1, V = 1.170906 V 14:57:21 Sensor Measurement, Sensor Number = 1, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 1.170906 (V, A, ohm, ...), Measured Value = 12.817935 (C, atm, Hz, ...) 14:57:21 Sensor Voltage Measurement, Sensor Number = 2, V = 0.001503 V 14:57:21 Sensor Measurement, Sensor Number = 2, Sensor Type = WATER_KEYENCE_FLOW Raw Sensor Output = 0.001503 (V, A, ohm, ...), Measured Value = 1.503260 (C, atm, Hz, ...) 14:57:21 Magnet Test Parameter Measurements P0 = HTRIM, Value = 24.99 P1 = Main (V), Value = 0.00 P2 = Core_Top, Value = 22.53 P3 = Top_Coil, Value = 24.65 P4 = H20 In, Value = 28.11 P5 = H20 Out, Value = 26.93 P6 = Ambient, Value = 17.42 P7 = Supply Pres(psi), Value = 122.93 P8 = Return Pres(psi), Value = 12.82 P9 = Flow Rate(gpm), Value = 1.50 14:57:22 Magnetic Field Measurement Measuring the magnet strength... 14:57:22 Integrated Voltage Measurement... 14:57:22 Magnet Current for HTRIM, Imag = 24.9911 A Vtrans = 2.082702 V, Factor V/A = 0.083333 V/A, Zero Current = 0.001333 A 14:57:31 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0200000 m VT = 0.001340787 Vs 14:57:38 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0200000 m VT = -0.001344471 Vs 14:57:39 Magnet Current for HTRIM, Imag = 24.9911 A Vtrans = 2.082706 V, Factor V/A = 0.083333 V/A, Zero Current = 0.001333 A 14:57:46 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0200000 m VT = 0.001339954 Vs 14:57:54 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0200000 m VT = -0.001344784 Vs 14:57:54 Magnet Current for HTRIM, Imag = 24.9911 A Vtrans = 2.082704 V, Factor V/A = 0.083333 V/A, Zero Current = 0.001333 A 14:58:02 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0200000 m VT = 0.001341129 Vs 14:58:09 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0200000 m VT = -0.001344925 Vs 14:58:10 Magnet Current for HTRIM, Imag = 24.9912 A Vtrans = 2.082712 V, Factor V/A = 0.083333 V/A, Zero Current = 0.001333 A 14:58:17 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0200000 m VT = 0.001340739 Vs 14:58:25 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0200000 m VT = -0.001344426 Vs 14:58:25 Magnet Current for HTRIM, Imag = 24.9910 A Vtrans = 2.082696 V, Factor V/A = 0.083333 V/A, Zero Current = 0.001333 A 14:58:33 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0200000 m VT = 0.001339351 Vs 14:58:40 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0200000 m VT = -0.001345089 Vs 14:58:40 Integrated Voltage Average and RMS of Measurements X0 = 0.0000000 m, Delta X = 0.0200000 m VT = 0.001342566 +- 0.000000274 Vs 14:58:40 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0000000 m, dx = 0.0200000 m, Nturns = 10.0 VT = 0.001342566 +- 0.000000274 Vs BL = 0.00671283 +- 0.00000137 Tm 14:58:44 Position moved: 0.000000 m 14:58:44 BL vs X Fit Coefficients: Cn sigCn n (Tm/m^n) ----- --------+-------- 0 0.00671 0.00000 Compute Fit Residuals: X0 BL sigBL BLfit BL-BLfit (m) (Tm) (Tm) (Tm) (Tm) ----------- --------+-------- -------- -------- 0.0000000 0.00671 0.00000 0.00671 0.00000 14:58:44 Magnet Strength Calculation Harmonic Number = 1 Number Of X0 Positions = 1 X0 BL sigBL (m) (Tm) (Tm) -------- --------+-------- 0.0000 0.00671 0.00000 SL = 0.0067128 +- 0.0000014 Tm^2/m^n 14:58:44 Magnet Strength Measurement Harmonic Number = 1 SL = 0.0067128 +- 0.0000014 Tm^2/m^n 14:58:47 Position moved: 0.000000 m 14:58:48 Magnet Current for HTRIM, Imag = 24.9911 A Vtrans = 2.082700 V, Factor V/A = 0.083333 V/A, Zero Current = 0.001333 A 14:58:48 Main (V) Voltage Measurement, V = 0.000002 V 14:58:52 Temperature Measurement, Probe Number = 0, T = 22.856000 C 14:58:53 Temperature Measurement, Probe Number = 1, T = 24.791000 C 14:58:53 Temperature Measurement, Probe Number = 2, T = 28.098000 C 14:58:54 Temperature Measurement, Probe Number = 3, T = 26.924000 C 14:58:55 Temperature Measurement, Probe Number = 4, T = 17.397000 C 14:58:55 Sensor Voltage Measurement, Sensor Number = 0, V = 2.642150 V 14:58:55 Sensor Measurement, Sensor Number = 0, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 2.642150 (V, A, ohm, ...), Measured Value = 123.161243 (C, atm, Hz, ...) 14:58:56 Sensor Voltage Measurement, Sensor Number = 1, V = 1.170701 V 14:58:56 Sensor Measurement, Sensor Number = 1, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 1.170701 (V, A, ohm, ...), Measured Value = 12.802575 (C, atm, Hz, ...) 14:58:57 Sensor Voltage Measurement, Sensor Number = 2, V = 0.001512 V 14:58:57 Sensor Measurement, Sensor Number = 2, Sensor Type = WATER_KEYENCE_FLOW Raw Sensor Output = 0.001512 (V, A, ohm, ...), Measured Value = 1.512473 (C, atm, Hz, ...) 14:58:57 Magnet Test Parameter Measurements P0 = HTRIM, Value = 24.99 P1 = Main (V), Value = 0.00 P2 = Core_Top, Value = 22.86 P3 = Top_Coil, Value = 24.79 P4 = H20 In, Value = 28.10 P5 = H20 Out, Value = 26.92 P6 = Ambient, Value = 17.40 P7 = Supply Pres(psi), Value = 123.16 P8 = Return Pres(psi), Value = 12.80 P9 = Flow Rate(gpm), Value = 1.51 14:58:57 Magnet system HTRIM current ramping to -25.0000 A... 14:59:38 Magnet Current for HTRIM, Imag = -25.0406 A Vtrans = -2.086609 V, Factor V/A = 0.083333 V/A, Zero Current = 0.001333 A 14:59:39 Main (V) Voltage Measurement, V = 0.000000 V 14:59:42 Temperature Measurement, Probe Number = 0, T = 22.754000 C 14:59:43 Temperature Measurement, Probe Number = 1, T = 24.810000 C 14:59:43 Temperature Measurement, Probe Number = 2, T = 28.097000 C 14:59:44 Temperature Measurement, Probe Number = 3, T = 26.917000 C 14:59:45 Temperature Measurement, Probe Number = 4, T = 17.326000 C 14:59:45 Sensor Voltage Measurement, Sensor Number = 0, V = 2.645913 V 14:59:45 Sensor Measurement, Sensor Number = 0, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 2.645913 (V, A, ohm, ...), Measured Value = 123.443505 (C, atm, Hz, ...) 14:59:46 Sensor Voltage Measurement, Sensor Number = 1, V = 1.171092 V 14:59:46 Sensor Measurement, Sensor Number = 1, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 1.171092 (V, A, ohm, ...), Measured Value = 12.831885 (C, atm, Hz, ...) 14:59:47 Sensor Voltage Measurement, Sensor Number = 2, V = 0.001523 V 14:59:47 Sensor Measurement, Sensor Number = 2, Sensor Type = WATER_KEYENCE_FLOW Raw Sensor Output = 0.001523 (V, A, ohm, ...), Measured Value = 1.523063 (C, atm, Hz, ...) 14:59:47 Magnet Test Parameter Measurements P0 = HTRIM, Value = -25.04 P1 = Main (V), Value = 0.00 P2 = Core_Top, Value = 22.75 P3 = Top_Coil, Value = 24.81 P4 = H20 In, Value = 28.10 P5 = H20 Out, Value = 26.92 P6 = Ambient, Value = 17.33 P7 = Supply Pres(psi), Value = 123.44 P8 = Return Pres(psi), Value = 12.83 P9 = Flow Rate(gpm), Value = 1.52 14:59:47 Magnetic Field Measurement Measuring the magnet strength... 14:59:47 Integrated Voltage Measurement... 14:59:48 Magnet Current for HTRIM, Imag = -25.0402 A Vtrans = -2.086573 V, Factor V/A = 0.083333 V/A, Zero Current = 0.001333 A 14:59:56 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0200000 m VT = -0.001348138 Vs 15:00:03 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0200000 m VT = 0.001340819 Vs 15:00:04 Magnet Current for HTRIM, Imag = -25.0399 A Vtrans = -2.086547 V, Factor V/A = 0.083333 V/A, Zero Current = 0.001333 A 15:00:11 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0200000 m VT = -0.001346797 Vs 15:00:19 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0200000 m VT = 0.001340528 Vs 15:00:19 Magnet Current for HTRIM, Imag = -25.0407 A Vtrans = -2.086614 V, Factor V/A = 0.083333 V/A, Zero Current = 0.001333 A 15:00:27 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0200000 m VT = -0.001347487 Vs 15:00:34 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0200000 m VT = 0.001341276 Vs 15:00:35 Magnet Current for HTRIM, Imag = -25.0390 A Vtrans = -2.086470 V, Factor V/A = 0.083333 V/A, Zero Current = 0.001333 A 15:00:42 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0200000 m VT = -0.001348587 Vs 15:00:49 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0200000 m VT = 0.001342295 Vs 15:00:50 Magnet Current for HTRIM, Imag = -25.0389 A Vtrans = -2.086468 V, Factor V/A = 0.083333 V/A, Zero Current = 0.001333 A 15:00:57 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0200000 m VT = -0.001346573 Vs 15:01:04 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0200000 m VT = 0.001342627 Vs 15:01:05 Integrated Voltage Average and RMS of Measurements X0 = 0.0000000 m, Delta X = 0.0200000 m VT = -0.001344513 +- 0.000000568 Vs 15:01:05 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0000000 m, dx = 0.0200000 m, Nturns = 10.0 VT = -0.001344513 +- 0.000000568 Vs BL = -0.00672256 +- 0.00000284 Tm 15:01:09 Position moved: 0.000000 m 15:01:09 BL vs X Fit Coefficients: Cn sigCn n (Tm/m^n) ----- --------+-------- 0 -0.00672 0.00000 Compute Fit Residuals: X0 BL sigBL BLfit BL-BLfit (m) (Tm) (Tm) (Tm) (Tm) ----------- --------+-------- -------- -------- 0.0000000 -0.00672 0.00000 -0.00672 0.00000 15:01:09 Magnet Strength Calculation Harmonic Number = 1 Number Of X0 Positions = 1 X0 BL sigBL (m) (Tm) (Tm) -------- --------+-------- 0.0000 -0.00672 0.00000 SL = -0.0067226 +- 0.0000028 Tm^2/m^n 15:01:09 Magnet Strength Measurement Harmonic Number = 1 SL = -0.0067226 +- 0.0000028 Tm^2/m^n 15:01:12 Position moved: 0.000000 m 15:01:12 Magnet Current for HTRIM, Imag = -25.0433 A Vtrans = -2.086829 V, Factor V/A = 0.083333 V/A, Zero Current = 0.001333 A 15:01:13 Main (V) Voltage Measurement, V = 0.000001 V 15:01:17 Temperature Measurement, Probe Number = 0, T = 22.848000 C 15:01:17 Temperature Measurement, Probe Number = 1, T = 24.895000 C 15:01:18 Temperature Measurement, Probe Number = 2, T = 28.105000 C 15:01:18 Temperature Measurement, Probe Number = 3, T = 26.902000 C 15:01:19 Temperature Measurement, Probe Number = 4, T = 17.327000 C 15:01:20 Sensor Voltage Measurement, Sensor Number = 0, V = 2.651385 V 15:01:20 Sensor Measurement, Sensor Number = 0, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 2.651385 (V, A, ohm, ...), Measured Value = 123.853883 (C, atm, Hz, ...) 15:01:20 Sensor Voltage Measurement, Sensor Number = 1, V = 1.170924 V 15:01:21 Sensor Measurement, Sensor Number = 1, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 1.170924 (V, A, ohm, ...), Measured Value = 12.819337 (C, atm, Hz, ...) 15:01:21 Sensor Voltage Measurement, Sensor Number = 2, V = 0.001523 V 15:01:21 Sensor Measurement, Sensor Number = 2, Sensor Type = WATER_KEYENCE_FLOW Raw Sensor Output = 0.001523 (V, A, ohm, ...), Measured Value = 1.523061 (C, atm, Hz, ...) 15:01:21 Magnet Test Parameter Measurements P0 = HTRIM, Value = -25.04 P1 = Main (V), Value = 0.00 P2 = Core_Top, Value = 22.85 P3 = Top_Coil, Value = 24.89 P4 = H20 In, Value = 28.11 P5 = H20 Out, Value = 26.90 P6 = Ambient, Value = 17.33 P7 = Supply Pres(psi), Value = 123.85 P8 = Return Pres(psi), Value = 12.82 P9 = Flow Rate(gpm), Value = 1.52 15:01:21 Magnet system HTRIM current ramping to -25.0000 A... 15:01:42 Magnet Current for HTRIM, Imag = -25.0386 A Vtrans = -2.086438 V, Factor V/A = 0.083333 V/A, Zero Current = 0.001333 A 15:01:43 Main (V) Voltage Measurement, V = 0.000000 V 15:01:47 Temperature Measurement, Probe Number = 0, T = 22.758000 C 15:01:47 Temperature Measurement, Probe Number = 1, T = 24.885000 C 15:01:48 Temperature Measurement, Probe Number = 2, T = 28.107000 C 15:01:48 Temperature Measurement, Probe Number = 3, T = 26.901000 C 15:01:49 Temperature Measurement, Probe Number = 4, T = 17.318000 C 15:01:50 Sensor Voltage Measurement, Sensor Number = 0, V = 2.651255 V 15:01:50 Sensor Measurement, Sensor Number = 0, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 2.651255 (V, A, ohm, ...), Measured Value = 123.844140 (C, atm, Hz, ...) 15:01:50 Sensor Voltage Measurement, Sensor Number = 1, V = 1.171414 V 15:01:50 Sensor Measurement, Sensor Number = 1, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 1.171414 (V, A, ohm, ...), Measured Value = 12.856050 (C, atm, Hz, ...) 15:01:51 Sensor Voltage Measurement, Sensor Number = 2, V = 0.001512 V 15:01:51 Sensor Measurement, Sensor Number = 2, Sensor Type = WATER_KEYENCE_FLOW Raw Sensor Output = 0.001512 (V, A, ohm, ...), Measured Value = 1.512312 (C, atm, Hz, ...) 15:01:51 Magnet Test Parameter Measurements P0 = HTRIM, Value = -25.04 P1 = Main (V), Value = 0.00 P2 = Core_Top, Value = 22.76 P3 = Top_Coil, Value = 24.89 P4 = H20 In, Value = 28.11 P5 = H20 Out, Value = 26.90 P6 = Ambient, Value = 17.32 P7 = Supply Pres(psi), Value = 123.84 P8 = Return Pres(psi), Value = 12.86 P9 = Flow Rate(gpm), Value = 1.51 15:01:51 Magnetic Field Measurement Measuring the magnet strength... 15:01:51 Integrated Voltage Measurement... 15:01:52 Magnet Current for HTRIM, Imag = -25.0389 A Vtrans = -2.086462 V, Factor V/A = 0.083333 V/A, Zero Current = 0.001333 A 15:02:00 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0200000 m VT = -0.001346875 Vs 15:02:08 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0200000 m VT = 0.001342931 Vs 15:02:08 Magnet Current for HTRIM, Imag = -25.0434 A Vtrans = -2.086841 V, Factor V/A = 0.083333 V/A, Zero Current = 0.001333 A 15:02:16 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0200000 m VT = -0.001347476 Vs 15:02:23 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0200000 m VT = 0.001342959 Vs 15:02:24 Magnet Current for HTRIM, Imag = -25.0424 A Vtrans = -2.086756 V, Factor V/A = 0.083333 V/A, Zero Current = 0.001333 A 15:02:31 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0200000 m VT = -0.001346728 Vs 15:02:38 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0200000 m VT = 0.001342180 Vs 15:02:39 Magnet Current for HTRIM, Imag = -25.0376 A Vtrans = -2.086358 V, Factor V/A = 0.083333 V/A, Zero Current = 0.001333 A 15:02:46 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0200000 m VT = -0.001346633 Vs 15:02:53 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0200000 m VT = 0.001341873 Vs 15:02:54 Magnet Current for HTRIM, Imag = -25.0377 A Vtrans = -2.086368 V, Factor V/A = 0.083333 V/A, Zero Current = 0.001333 A 15:03:01 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0200000 m VT = -0.001345886 Vs 15:03:09 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0200000 m VT = 0.001342058 Vs 15:03:09 Integrated Voltage Average and RMS of Measurements X0 = 0.0000000 m, Delta X = 0.0200000 m VT = -0.001344560 +- 0.000000447 Vs 15:03:09 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0000000 m, dx = 0.0200000 m, Nturns = 10.0 VT = -0.001344560 +- 0.000000447 Vs BL = -0.00672280 +- 0.00000224 Tm 15:03:12 Position moved: 0.000000 m 15:03:13 BL vs X Fit Coefficients: Cn sigCn n (Tm/m^n) ----- --------+-------- 0 -0.00672 0.00000 Compute Fit Residuals: X0 BL sigBL BLfit BL-BLfit (m) (Tm) (Tm) (Tm) (Tm) ----------- --------+-------- -------- -------- 0.0000000 -0.00672 0.00000 -0.00672 0.00000 15:03:13 Magnet Strength Calculation Harmonic Number = 1 Number Of X0 Positions = 1 X0 BL sigBL (m) (Tm) (Tm) -------- --------+-------- 0.0000 -0.00672 0.00000 SL = -0.0067228 +- 0.0000022 Tm^2/m^n 15:03:13 Magnet Strength Measurement Harmonic Number = 1 SL = -0.0067228 +- 0.0000022 Tm^2/m^n 15:03:15 Position moved: 0.000000 m 15:03:16 Magnet Current for HTRIM, Imag = -25.0375 A Vtrans = -2.086348 V, Factor V/A = 0.083333 V/A, Zero Current = 0.001333 A 15:03:17 Main (V) Voltage Measurement, V = 0.000001 V 15:03:21 Temperature Measurement, Probe Number = 0, T = 22.638000 C 15:03:21 Temperature Measurement, Probe Number = 1, T = 24.850000 C 15:03:22 Temperature Measurement, Probe Number = 2, T = 28.110000 C 15:03:22 Temperature Measurement, Probe Number = 3, T = 26.902000 C 15:03:23 Temperature Measurement, Probe Number = 4, T = 17.324000 C 15:03:24 Sensor Voltage Measurement, Sensor Number = 0, V = 2.641208 V 15:03:24 Sensor Measurement, Sensor Number = 0, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 2.641208 (V, A, ohm, ...), Measured Value = 123.090630 (C, atm, Hz, ...) 15:03:24 Sensor Voltage Measurement, Sensor Number = 1, V = 1.171112 V 15:03:24 Sensor Measurement, Sensor Number = 1, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 1.171112 (V, A, ohm, ...), Measured Value = 12.833385 (C, atm, Hz, ...) 15:03:25 Sensor Voltage Measurement, Sensor Number = 2, V = 0.001523 V 15:03:25 Sensor Measurement, Sensor Number = 2, Sensor Type = WATER_KEYENCE_FLOW Raw Sensor Output = 0.001523 (V, A, ohm, ...), Measured Value = 1.522614 (C, atm, Hz, ...) 15:03:25 Magnet Test Parameter Measurements P0 = HTRIM, Value = -25.04 P1 = Main (V), Value = 0.00 P2 = Core_Top, Value = 22.64 P3 = Top_Coil, Value = 24.85 P4 = H20 In, Value = 28.11 P5 = H20 Out, Value = 26.90 P6 = Ambient, Value = 17.32 P7 = Supply Pres(psi), Value = 123.09 P8 = Return Pres(psi), Value = 12.83 P9 = Flow Rate(gpm), Value = 1.52 15:03:25 Magnet system HTRIM current ramping to -25.0000 A... 15:03:46 Magnet Current for HTRIM, Imag = -25.0385 A Vtrans = -2.086429 V, Factor V/A = 0.083333 V/A, Zero Current = 0.001333 A 15:03:47 Main (V) Voltage Measurement, V = 0.000001 V 15:03:50 Temperature Measurement, Probe Number = 0, T = 22.756000 C 15:03:51 Temperature Measurement, Probe Number = 1, T = 24.840000 C 15:03:52 Temperature Measurement, Probe Number = 2, T = 28.108000 C 15:03:52 Temperature Measurement, Probe Number = 3, T = 26.902000 C 15:03:53 Temperature Measurement, Probe Number = 4, T = 17.345000 C 15:03:53 Sensor Voltage Measurement, Sensor Number = 0, V = 2.647333 V 15:03:53 Sensor Measurement, Sensor Number = 0, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 2.647333 (V, A, ohm, ...), Measured Value = 123.549990 (C, atm, Hz, ...) 15:03:54 Sensor Voltage Measurement, Sensor Number = 1, V = 1.171230 V 15:03:54 Sensor Measurement, Sensor Number = 1, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 1.171230 (V, A, ohm, ...), Measured Value = 12.842280 (C, atm, Hz, ...) 15:03:55 Sensor Voltage Measurement, Sensor Number = 2, V = 0.001513 V 15:03:55 Sensor Measurement, Sensor Number = 2, Sensor Type = WATER_KEYENCE_FLOW Raw Sensor Output = 0.001513 (V, A, ohm, ...), Measured Value = 1.512531 (C, atm, Hz, ...) 15:03:55 Magnet Test Parameter Measurements P0 = HTRIM, Value = -25.04 P1 = Main (V), Value = 0.00 P2 = Core_Top, Value = 22.76 P3 = Top_Coil, Value = 24.84 P4 = H20 In, Value = 28.11 P5 = H20 Out, Value = 26.90 P6 = Ambient, Value = 17.34 P7 = Supply Pres(psi), Value = 123.55 P8 = Return Pres(psi), Value = 12.84 P9 = Flow Rate(gpm), Value = 1.51 15:03:55 Magnetic Field Measurement Measuring the magnet strength... 15:03:55 Integrated Voltage Measurement... 15:03:55 Magnet Current for HTRIM, Imag = -25.0413 A Vtrans = -2.086663 V, Factor V/A = 0.083333 V/A, Zero Current = 0.001333 A 15:04:04 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0200000 m VT = -0.001345928 Vs 15:04:11 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0200000 m VT = 0.001343406 Vs 15:04:12 Magnet Current for HTRIM, Imag = -25.0431 A Vtrans = -2.086814 V, Factor V/A = 0.083333 V/A, Zero Current = 0.001333 A 15:04:19 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0200000 m VT = -0.001346901 Vs 15:04:26 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0200000 m VT = 0.001341420 Vs 15:04:27 Magnet Current for HTRIM, Imag = -25.0425 A Vtrans = -2.086766 V, Factor V/A = 0.083333 V/A, Zero Current = 0.001333 A 15:04:34 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0200000 m VT = -0.001347643 Vs 15:04:41 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0200000 m VT = 0.001342279 Vs 15:04:42 Magnet Current for HTRIM, Imag = -25.0417 A Vtrans = -2.086696 V, Factor V/A = 0.083333 V/A, Zero Current = 0.001333 A 15:04:50 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0200000 m VT = -0.001348140 Vs 15:04:57 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0200000 m VT = 0.001341730 Vs 15:04:58 Magnet Current for HTRIM, Imag = -25.0388 A Vtrans = -2.086453 V, Factor V/A = 0.083333 V/A, Zero Current = 0.001333 A 15:05:05 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0200000 m VT = -0.001347501 Vs 15:05:12 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0200000 m VT = 0.001341671 Vs 15:05:12 Integrated Voltage Average and RMS of Measurements X0 = 0.0000000 m, Delta X = 0.0200000 m VT = -0.001344662 +- 0.000000290 Vs 15:05:12 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0000000 m, dx = 0.0200000 m, Nturns = 10.0 VT = -0.001344662 +- 0.000000290 Vs BL = -0.00672331 +- 0.00000145 Tm 15:05:16 Position moved: 0.000000 m 15:05:16 BL vs X Fit Coefficients: Cn sigCn n (Tm/m^n) ----- --------+-------- 0 -0.00672 0.00000 Compute Fit Residuals: X0 BL sigBL BLfit BL-BLfit (m) (Tm) (Tm) (Tm) (Tm) ----------- --------+-------- -------- -------- 0.0000000 -0.00672 0.00000 -0.00672 0.00000 15:05:16 Magnet Strength Calculation Harmonic Number = 1 Number Of X0 Positions = 1 X0 BL sigBL (m) (Tm) (Tm) -------- --------+-------- 0.0000 -0.00672 0.00000 SL = -0.0067233 +- 0.0000015 Tm^2/m^n 15:05:16 Magnet Strength Measurement Harmonic Number = 1 SL = -0.0067233 +- 0.0000015 Tm^2/m^n 15:05:19 Position moved: 0.000000 m 15:05:20 Magnet Current for HTRIM, Imag = -25.0390 A Vtrans = -2.086469 V, Factor V/A = 0.083333 V/A, Zero Current = 0.001333 A 15:05:20 Main (V) Voltage Measurement, V = 0.000001 V 15:05:24 Temperature Measurement, Probe Number = 0, T = 22.981000 C 15:05:24 Temperature Measurement, Probe Number = 1, T = 24.958000 C 15:05:25 Temperature Measurement, Probe Number = 2, T = 28.108000 C 15:05:26 Temperature Measurement, Probe Number = 3, T = 26.905000 C 15:05:26 Temperature Measurement, Probe Number = 4, T = 17.306000 C 15:05:27 Sensor Voltage Measurement, Sensor Number = 0, V = 2.644406 V 15:05:27 Sensor Measurement, Sensor Number = 0, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 2.644406 (V, A, ohm, ...), Measured Value = 123.330472 (C, atm, Hz, ...) 15:05:28 Sensor Voltage Measurement, Sensor Number = 1, V = 1.170725 V 15:05:28 Sensor Measurement, Sensor Number = 1, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 1.170725 (V, A, ohm, ...), Measured Value = 12.804353 (C, atm, Hz, ...) 15:05:28 Sensor Voltage Measurement, Sensor Number = 2, V = 0.001512 V 15:05:28 Sensor Measurement, Sensor Number = 2, Sensor Type = WATER_KEYENCE_FLOW Raw Sensor Output = 0.001512 (V, A, ohm, ...), Measured Value = 1.512487 (C, atm, Hz, ...) 15:05:28 Magnet Test Parameter Measurements P0 = HTRIM, Value = -25.04 P1 = Main (V), Value = 0.00 P2 = Core_Top, Value = 22.98 P3 = Top_Coil, Value = 24.96 P4 = H20 In, Value = 28.11 P5 = H20 Out, Value = 26.91 P6 = Ambient, Value = 17.31 P7 = Supply Pres(psi), Value = 123.33 P8 = Return Pres(psi), Value = 12.80 P9 = Flow Rate(gpm), Value = 1.51 15:05:31 Position moved: 0.000000 m 15:05:34 Magnet system HTRIM current ramping to 0.0000 A...