Coil Constant for 0_625DC25q_coiladd using magnet Q1E#1 New Calibrated Coil Constant = 1769.62071, File coilpar.ru (Old) Coil Constant = 1763.42345 New Constant/Old Constant = 1.0035, Old Coil Radius = 0.0075200, New Radius = 0.0075173, Number of coils = 2 (Only change radius if Number of coils = 1!) Wire Strength = 0.2778 T/A, Coil Strength = 0.2769 T/A, Magnet Type is Quadrupole at Imag = 50.026 for coil. Number of Coil Turns = 10 Coil Constant Calculated from radius, turns and magnet order = 1768.3341 (Should be within 5% of Calibrated Coil Constant)