SLAC Magnetic Measurements Date: 08-13-2010 Time: 10:25:25 Project: FACET_Sector_20 Magnet Type: Dipole Magnet Name: WIG2E Serial Number: 002153 Run Number: 9 Measurement Device: 10 WIRE Operator: sda Comment: Integral Bdl vs x at 5 amps. Normal Stdz, then 0 amps then 5 amps Standardization for Main: 3 cycles from 10.0 A to 150.0 A to 10.0 A Test Currents for Main (A): 0.00 5.00 Integrated Field Strength Distance the wire was to be moved = 0.010000 m Number of turns in the wire = 10 Number of measurements for averaging = 6 Main +/- sigma BL sigBL BL/I sigBL/I (A) (A) (Tm) (Tm) (Tm/A) (Tm/A) ----------+---------- ------------+------------ -------------+------------- 4.9742 0.0001 -0.017074 7.14459e-06 -0.003432519 1.436332e-06