SLAC Magnetic Measurements Date: 01-23-2012 Time: 15:51:01 Project: FACET_Sector_10 Magnet Type: Dipole Magnet Name: B104EP Serial Number: 1.071.4T Run Number: 5 Measurement Device: 10_wire Operator: sda Comment: Integrated strength. Main ps off after stdz and ramping to 859 A. Trim scan from -0. to -0.5 A Test Currents for Main (A): 0.00 0.00 0.00 0.00 0.00 0.00 0.00 0.00 0.00 0.00 0.00 Test Currents for Trim (A): 0.00 -0.05 -0.10 -0.15 -0.20 -0.25 -0.30 -0.35 -0.40 -0.45 -0.50 15:51:28 Magnet Current for Main, Imag = -0.0093 A Vtrans = -0.000093 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 15:51:28 Magnet Current for Main, Imag = -0.0093 A Vtrans = -0.000093 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 15:51:29 Magnet Current for Main, Imag = -0.0093 A Vtrans = -0.000093 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 15:51:29 Magnet Current for Main, Imag = -0.0094 A Vtrans = -0.000094 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 15:51:29 Magnet Avgerage Current Measurement for Main : Imag = -0.0093 +- 0.0000 A 15:51:31 Initial Zero Current Measurement for Main, Imag_zero = -0.00934203 A 15:51:31 Initial Zero Current Measurement for Trim, Imag_zero = 0 A 15:51:43 Name = None, is zeroed 15:51:43 Magnet system Main current ramping to 0.0000 A... 15:52:13 Magnet system Trim current ramping to 0.0000 A... 15:52:43 Magnetic Field Measurement Measuring the magnet strength... 15:52:43 Integrated Voltage Measurement... 15:52:44 Magnet Current for Main, Imag = -0.0001 A Vtrans = -0.000094 V, Factor V/A = 0.010000 V/A, Zero Current = -0.009342 A 15:52:44 Magnet Current for Trim, Imag = -0.0002 A Vtrans = -0.000013 V, Factor V/A = 0.083333 V/A, Zero Current = 0.000000 A 15:53:02 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.000411755 Vs 15:53:19 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.000432009 Vs 15:53:20 Magnet Current for Main, Imag = 0.0001 A Vtrans = -0.000093 V, Factor V/A = 0.010000 V/A, Zero Current = -0.009342 A 15:53:21 Magnet Current for Trim, Imag = -0.0002 A Vtrans = -0.000014 V, Factor V/A = 0.083333 V/A, Zero Current = 0.000000 A 15:53:38 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.000411682 Vs 15:53:55 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.000432928 Vs 15:53:55 Magnet Current for Main, Imag = -0.0000 A Vtrans = -0.000093 V, Factor V/A = 0.010000 V/A, Zero Current = -0.009342 A 15:53:56 Magnet Current for Trim, Imag = -0.0002 A Vtrans = -0.000013 V, Factor V/A = 0.083333 V/A, Zero Current = 0.000000 A 15:54:13 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.000410993 Vs 15:54:30 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.000432777 Vs 15:54:31 Magnet Current for Main, Imag = -0.0000 A Vtrans = -0.000094 V, Factor V/A = 0.010000 V/A, Zero Current = -0.009342 A 15:54:32 Magnet Current for Trim, Imag = -0.0002 A Vtrans = -0.000014 V, Factor V/A = 0.083333 V/A, Zero Current = 0.000000 A 15:54:49 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.000410704 Vs 15:55:06 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.000434115 Vs 15:55:06 Integrated Voltage Average and RMS of Measurements X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.000422120 +- 0.000000240 Vs 15:55:06 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0000000 m, dx = 0.0100000 m, Nturns = 10.0 VT = 0.000422120 +- 0.000000240 Vs BL = 0.00422120 +- 0.00000240 Tm 15:55:14 BL vs X Fit Coefficients: Cn sigCn n (Tm/m^n) ----- --------+-------- 0 0.00422 0.00000 Compute Fit Residuals: X0 BL sigBL BLfit BL-BLfit (m) (Tm) (Tm) (Tm) (Tm) ----------- --------+-------- -------- -------- 0.0000000 0.00422 0.00000 0.00422 0.00000 15:55:14 Magnet Strength Calculation Harmonic Number = 1 Number Of X0 Positions = 1 X0 BL sigBL (m) (Tm) (Tm) -------- --------+-------- 0.0000 0.00422 0.00000 SL = 0.0042212 +- 0.0000024 Tm^2/m^n 15:55:14 Magnet Strength Measurement Harmonic Number = 1 SL = 0.0042212 +- 0.0000024 Tm^2/m^n 15:55:14 Magnet system Main current ramping to 0.0000 A... 15:55:44 Magnet system Trim current ramping to -0.0500 A... 15:56:14 Magnetic Field Measurement Measuring the magnet strength... 15:56:14 Integrated Voltage Measurement... 15:56:15 Magnet Current for Main, Imag = -0.0000 A Vtrans = -0.000094 V, Factor V/A = 0.010000 V/A, Zero Current = -0.009342 A 15:56:16 Magnet Current for Trim, Imag = -0.0483 A Vtrans = -0.004026 V, Factor V/A = 0.083333 V/A, Zero Current = 0.000000 A 15:56:34 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.000339532 Vs 15:56:51 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.000359216 Vs 15:56:51 Magnet Current for Main, Imag = -0.0001 A Vtrans = -0.000094 V, Factor V/A = 0.010000 V/A, Zero Current = -0.009342 A 15:56:52 Magnet Current for Trim, Imag = -0.0483 A Vtrans = -0.004027 V, Factor V/A = 0.083333 V/A, Zero Current = 0.000000 A 15:57:09 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.000337602 Vs 15:57:26 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.000360731 Vs 15:57:27 Magnet Current for Main, Imag = -0.0001 A Vtrans = -0.000095 V, Factor V/A = 0.010000 V/A, Zero Current = -0.009342 A 15:57:27 Magnet Current for Trim, Imag = -0.0483 A Vtrans = -0.004028 V, Factor V/A = 0.083333 V/A, Zero Current = 0.000000 A 15:57:44 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.000336290 Vs 15:58:02 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.000360769 Vs 15:58:02 Magnet Current for Main, Imag = -0.0001 A Vtrans = -0.000094 V, Factor V/A = 0.010000 V/A, Zero Current = -0.009342 A 15:58:03 Magnet Current for Trim, Imag = -0.0483 A Vtrans = -0.004027 V, Factor V/A = 0.083333 V/A, Zero Current = 0.000000 A 15:58:20 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.000335728 Vs 15:58:37 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.000359620 Vs 15:58:37 Integrated Voltage Average and RMS of Measurements X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.000348686 +- 0.000000662 Vs 15:58:37 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0000000 m, dx = 0.0100000 m, Nturns = 10.0 VT = 0.000348686 +- 0.000000662 Vs BL = 0.00348686 +- 0.00000662 Tm 15:58:45 BL vs X Fit Coefficients: Cn sigCn n (Tm/m^n) ----- --------+-------- 0 0.00349 0.00001 Compute Fit Residuals: X0 BL sigBL BLfit BL-BLfit (m) (Tm) (Tm) (Tm) (Tm) ----------- --------+-------- -------- -------- 0.0000000 0.00349 0.00001 0.00349 0.00000 15:58:45 Magnet Strength Calculation Harmonic Number = 1 Number Of X0 Positions = 1 X0 BL sigBL (m) (Tm) (Tm) -------- --------+-------- 0.0000 0.00349 0.00001 SL = 0.0034869 +- 0.0000066 Tm^2/m^n 15:58:45 Magnet Strength Measurement Harmonic Number = 1 SL = 0.0034869 +- 0.0000066 Tm^2/m^n 15:58:45 Magnet system Main current ramping to 0.0000 A... 15:59:15 Magnet system Trim current ramping to -0.1000 A... 15:59:46 Magnetic Field Measurement Measuring the magnet strength... 15:59:46 Integrated Voltage Measurement... 15:59:46 Magnet Current for Main, Imag = -0.0001 A Vtrans = -0.000094 V, Factor V/A = 0.010000 V/A, Zero Current = -0.009342 A 15:59:47 Magnet Current for Trim, Imag = -0.0996 A Vtrans = -0.008296 V, Factor V/A = 0.083333 V/A, Zero Current = 0.000000 A 16:00:05 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.000261607 Vs 16:00:22 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.000277429 Vs 16:00:22 Magnet Current for Main, Imag = -0.0001 A Vtrans = -0.000095 V, Factor V/A = 0.010000 V/A, Zero Current = -0.009342 A 16:00:23 Magnet Current for Trim, Imag = -0.0995 A Vtrans = -0.008295 V, Factor V/A = 0.083333 V/A, Zero Current = 0.000000 A 16:00:40 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.000261246 Vs 16:00:57 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.000277135 Vs 16:00:58 Magnet Current for Main, Imag = -0.0001 A Vtrans = -0.000094 V, Factor V/A = 0.010000 V/A, Zero Current = -0.009342 A 16:00:59 Magnet Current for Trim, Imag = -0.0996 A Vtrans = -0.008296 V, Factor V/A = 0.083333 V/A, Zero Current = 0.000000 A 16:01:16 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.000260028 Vs 16:01:33 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.000278684 Vs 16:01:33 Magnet Current for Main, Imag = -0.0001 A Vtrans = -0.000094 V, Factor V/A = 0.010000 V/A, Zero Current = -0.009342 A 16:01:34 Magnet Current for Trim, Imag = -0.0996 A Vtrans = -0.008297 V, Factor V/A = 0.083333 V/A, Zero Current = 0.000000 A 16:01:51 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.000258076 Vs 16:02:08 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.000279260 Vs 16:02:08 Integrated Voltage Average and RMS of Measurements X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.000269183 +- 0.000000319 Vs 16:02:08 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0000000 m, dx = 0.0100000 m, Nturns = 10.0 VT = 0.000269183 +- 0.000000319 Vs BL = 0.00269183 +- 0.00000319 Tm 16:02:16 BL vs X Fit Coefficients: Cn sigCn n (Tm/m^n) ----- --------+-------- 0 0.00269 0.00000 Compute Fit Residuals: X0 BL sigBL BLfit BL-BLfit (m) (Tm) (Tm) (Tm) (Tm) ----------- --------+-------- -------- -------- 0.0000000 0.00269 0.00000 0.00269 0.00000 16:02:16 Magnet Strength Calculation Harmonic Number = 1 Number Of X0 Positions = 1 X0 BL sigBL (m) (Tm) (Tm) -------- --------+-------- 0.0000 0.00269 0.00000 SL = 0.0026918 +- 0.0000032 Tm^2/m^n 16:02:16 Magnet Strength Measurement Harmonic Number = 1 SL = 0.0026918 +- 0.0000032 Tm^2/m^n 16:02:16 Magnet system Main current ramping to 0.0000 A... 16:02:47 Magnet system Trim current ramping to -0.1500 A... 16:03:17 Magnetic Field Measurement Measuring the magnet strength... 16:03:17 Integrated Voltage Measurement... 16:03:17 Magnet Current for Main, Imag = -0.0000 A Vtrans = -0.000094 V, Factor V/A = 0.010000 V/A, Zero Current = -0.009342 A 16:03:18 Magnet Current for Trim, Imag = -0.1508 A Vtrans = -0.012566 V, Factor V/A = 0.083333 V/A, Zero Current = 0.000000 A 16:03:36 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.000178913 Vs 16:03:53 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.000201508 Vs 16:03:54 Magnet Current for Main, Imag = -0.0001 A Vtrans = -0.000094 V, Factor V/A = 0.010000 V/A, Zero Current = -0.009342 A 16:03:54 Magnet Current for Trim, Imag = -0.1508 A Vtrans = -0.012566 V, Factor V/A = 0.083333 V/A, Zero Current = 0.000000 A 16:04:11 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.000178285 Vs 16:04:29 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.000199740 Vs 16:04:29 Magnet Current for Main, Imag = -0.0001 A Vtrans = -0.000095 V, Factor V/A = 0.010000 V/A, Zero Current = -0.009342 A 16:04:30 Magnet Current for Trim, Imag = -0.1508 A Vtrans = -0.012566 V, Factor V/A = 0.083333 V/A, Zero Current = 0.000000 A 16:04:47 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.000177618 Vs 16:05:04 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.000201067 Vs 16:05:05 Magnet Current for Main, Imag = -0.0001 A Vtrans = -0.000094 V, Factor V/A = 0.010000 V/A, Zero Current = -0.009342 A 16:05:05 Magnet Current for Trim, Imag = -0.1508 A Vtrans = -0.012567 V, Factor V/A = 0.083333 V/A, Zero Current = 0.000000 A 16:05:22 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.000176108 Vs 16:05:39 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.000200729 Vs 16:05:39 Integrated Voltage Average and RMS of Measurements X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.000189246 +- 0.000000648 Vs 16:05:39 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0000000 m, dx = 0.0100000 m, Nturns = 10.0 VT = 0.000189246 +- 0.000000648 Vs BL = 0.00189246 +- 0.00000648 Tm 16:05:47 BL vs X Fit Coefficients: Cn sigCn n (Tm/m^n) ----- --------+-------- 0 0.00189 0.00001 Compute Fit Residuals: X0 BL sigBL BLfit BL-BLfit (m) (Tm) (Tm) (Tm) (Tm) ----------- --------+-------- -------- -------- 0.0000000 0.00189 0.00001 0.00189 0.00000 16:05:47 Magnet Strength Calculation Harmonic Number = 1 Number Of X0 Positions = 1 X0 BL sigBL (m) (Tm) (Tm) -------- --------+-------- 0.0000 0.00189 0.00001 SL = 0.0018925 +- 0.0000065 Tm^2/m^n 16:05:47 Magnet Strength Measurement Harmonic Number = 1 SL = 0.0018925 +- 0.0000065 Tm^2/m^n 16:05:47 Magnet system Main current ramping to 0.0000 A... 16:06:18 Magnet system Trim current ramping to -0.2000 A... 16:06:48 Magnetic Field Measurement Measuring the magnet strength... 16:06:48 Integrated Voltage Measurement... 16:06:49 Magnet Current for Main, Imag = -0.0001 A Vtrans = -0.000094 V, Factor V/A = 0.010000 V/A, Zero Current = -0.009342 A 16:06:49 Magnet Current for Trim, Imag = -0.1986 A Vtrans = -0.016551 V, Factor V/A = 0.083333 V/A, Zero Current = 0.000000 A 16:07:07 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.000104365 Vs 16:07:24 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.000126805 Vs 16:07:25 Magnet Current for Main, Imag = -0.0001 A Vtrans = -0.000095 V, Factor V/A = 0.010000 V/A, Zero Current = -0.009342 A 16:07:25 Magnet Current for Trim, Imag = -0.1986 A Vtrans = -0.016551 V, Factor V/A = 0.083333 V/A, Zero Current = 0.000000 A 16:07:43 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.000102979 Vs 16:08:00 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.000126652 Vs 16:08:00 Magnet Current for Main, Imag = -0.0001 A Vtrans = -0.000094 V, Factor V/A = 0.010000 V/A, Zero Current = -0.009342 A 16:08:01 Magnet Current for Trim, Imag = -0.1986 A Vtrans = -0.016550 V, Factor V/A = 0.083333 V/A, Zero Current = 0.000000 A 16:08:18 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.000102217 Vs 16:08:35 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.000126280 Vs 16:08:36 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000093 V, Factor V/A = 0.010000 V/A, Zero Current = -0.009342 A 16:08:37 Magnet Current for Trim, Imag = -0.1986 A Vtrans = -0.016551 V, Factor V/A = 0.083333 V/A, Zero Current = 0.000000 A 16:08:54 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.000102856 Vs 16:09:11 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.000125151 Vs 16:09:11 Integrated Voltage Average and RMS of Measurements X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.000114663 +- 0.000000608 Vs 16:09:11 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0000000 m, dx = 0.0100000 m, Nturns = 10.0 VT = 0.000114663 +- 0.000000608 Vs BL = 0.00114663 +- 0.00000608 Tm 16:09:19 BL vs X Fit Coefficients: Cn sigCn n (Tm/m^n) ----- --------+-------- 0 0.00115 0.00001 Compute Fit Residuals: X0 BL sigBL BLfit BL-BLfit (m) (Tm) (Tm) (Tm) (Tm) ----------- --------+-------- -------- -------- 0.0000000 0.00115 0.00001 0.00115 0.00000 16:09:19 Magnet Strength Calculation Harmonic Number = 1 Number Of X0 Positions = 1 X0 BL sigBL (m) (Tm) (Tm) -------- --------+-------- 0.0000 0.00115 0.00001 SL = 0.0011466 +- 0.0000061 Tm^2/m^n 16:09:19 Magnet Strength Measurement Harmonic Number = 1 SL = 0.0011466 +- 0.0000061 Tm^2/m^n 16:09:19 Magnet system Main current ramping to 0.0000 A... 16:09:49 Magnet system Trim current ramping to -0.2500 A... 16:10:19 Magnetic Field Measurement Measuring the magnet strength... 16:10:19 Integrated Voltage Measurement... 16:10:20 Magnet Current for Main, Imag = -0.0001 A Vtrans = -0.000094 V, Factor V/A = 0.010000 V/A, Zero Current = -0.009342 A 16:10:20 Magnet Current for Trim, Imag = -0.2499 A Vtrans = -0.020826 V, Factor V/A = 0.083333 V/A, Zero Current = 0.000000 A 16:10:38 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.000023952 Vs 16:10:55 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.000047105 Vs 16:10:56 Magnet Current for Main, Imag = -0.0001 A Vtrans = -0.000094 V, Factor V/A = 0.010000 V/A, Zero Current = -0.009342 A 16:10:57 Magnet Current for Trim, Imag = -0.2499 A Vtrans = -0.020826 V, Factor V/A = 0.083333 V/A, Zero Current = 0.000000 A 16:11:14 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.000023268 Vs 16:11:31 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.000045728 Vs 16:11:31 Magnet Current for Main, Imag = -0.0001 A Vtrans = -0.000094 V, Factor V/A = 0.010000 V/A, Zero Current = -0.009342 A 16:11:32 Magnet Current for Trim, Imag = -0.2499 A Vtrans = -0.020826 V, Factor V/A = 0.083333 V/A, Zero Current = 0.000000 A 16:11:49 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.000023798 Vs 16:12:06 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.000044983 Vs 16:12:07 Magnet Current for Main, Imag = -0.0001 A Vtrans = -0.000095 V, Factor V/A = 0.010000 V/A, Zero Current = -0.009342 A 16:12:08 Magnet Current for Trim, Imag = -0.2499 A Vtrans = -0.020827 V, Factor V/A = 0.083333 V/A, Zero Current = 0.000000 A 16:12:25 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.000023475 Vs 16:12:42 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.000044840 Vs 16:12:42 Integrated Voltage Average and RMS of Measurements X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.000034644 +- 0.000000526 Vs 16:12:42 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0000000 m, dx = 0.0100000 m, Nturns = 10.0 VT = 0.000034644 +- 0.000000526 Vs BL = 0.00034644 +- 0.00000526 Tm 16:12:50 BL vs X Fit Coefficients: Cn sigCn n (Tm/m^n) ----- --------+-------- 0 0.00035 0.00001 Compute Fit Residuals: X0 BL sigBL BLfit BL-BLfit (m) (Tm) (Tm) (Tm) (Tm) ----------- --------+-------- -------- -------- 0.0000000 0.00035 0.00001 0.00035 0.00000 16:12:50 Magnet Strength Calculation Harmonic Number = 1 Number Of X0 Positions = 1 X0 BL sigBL (m) (Tm) (Tm) -------- --------+-------- 0.0000 0.00035 0.00001 SL = 0.0003464 +- 0.0000053 Tm^2/m^n 16:12:50 Magnet Strength Measurement Harmonic Number = 1 SL = 0.0003464 +- 0.0000053 Tm^2/m^n 16:12:50 Magnet system Main current ramping to 0.0000 A... 16:13:20 Magnet system Trim current ramping to -0.3000 A... 16:13:50 Magnetic Field Measurement Measuring the magnet strength... 16:13:50 Integrated Voltage Measurement... 16:13:51 Magnet Current for Main, Imag = -0.0001 A Vtrans = -0.000094 V, Factor V/A = 0.010000 V/A, Zero Current = -0.009342 A 16:13:51 Magnet Current for Trim, Imag = -0.3010 A Vtrans = -0.025083 V, Factor V/A = 0.083333 V/A, Zero Current = 0.000000 A 16:14:09 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = -0.000054684 Vs 16:14:26 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = 0.000033720 Vs 16:14:27 Magnet Current for Main, Imag = -0.0000 A Vtrans = -0.000094 V, Factor V/A = 0.010000 V/A, Zero Current = -0.009342 A 16:14:28 Magnet Current for Trim, Imag = -0.3010 A Vtrans = -0.025082 V, Factor V/A = 0.083333 V/A, Zero Current = 0.000000 A 16:14:45 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = -0.000056996 Vs 16:15:02 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = 0.000034318 Vs 16:15:02 Magnet Current for Main, Imag = -0.0001 A Vtrans = -0.000094 V, Factor V/A = 0.010000 V/A, Zero Current = -0.009342 A 16:15:03 Magnet Current for Trim, Imag = -0.3010 A Vtrans = -0.025082 V, Factor V/A = 0.083333 V/A, Zero Current = 0.000000 A 16:15:20 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = -0.000055811 Vs 16:15:37 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = 0.000034440 Vs 16:15:38 Magnet Current for Main, Imag = -0.0001 A Vtrans = -0.000094 V, Factor V/A = 0.010000 V/A, Zero Current = -0.009342 A 16:15:38 Magnet Current for Trim, Imag = -0.3010 A Vtrans = -0.025082 V, Factor V/A = 0.083333 V/A, Zero Current = 0.000000 A 16:15:56 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = -0.000057351 Vs 16:16:13 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = 0.000033898 Vs 16:16:13 Integrated Voltage Average and RMS of Measurements X0 = 0.0000000 m, Delta X = 0.0100000 m VT = -0.000045152 +- 0.000000588 Vs 16:16:13 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0000000 m, dx = 0.0100000 m, Nturns = 10.0 VT = -0.000045152 +- 0.000000588 Vs BL = -0.00045152 +- 0.00000588 Tm 16:16:21 BL vs X Fit Coefficients: Cn sigCn n (Tm/m^n) ----- --------+-------- 0 -0.00045 0.00001 Compute Fit Residuals: X0 BL sigBL BLfit BL-BLfit (m) (Tm) (Tm) (Tm) (Tm) ----------- --------+-------- -------- -------- 0.0000000 -0.00045 0.00001 -0.00045 0.00000 16:16:21 Magnet Strength Calculation Harmonic Number = 1 Number Of X0 Positions = 1 X0 BL sigBL (m) (Tm) (Tm) -------- --------+-------- 0.0000 -0.00045 0.00001 SL = -0.0004515 +- 0.0000059 Tm^2/m^n 16:16:21 Magnet Strength Measurement Harmonic Number = 1 SL = -0.0004515 +- 0.0000059 Tm^2/m^n 16:16:21 Magnet system Main current ramping to 0.0000 A... 16:16:51 Magnet system Trim current ramping to -0.3500 A... 16:17:21 Magnetic Field Measurement Measuring the magnet strength... 16:17:21 Integrated Voltage Measurement... 16:17:22 Magnet Current for Main, Imag = -0.0001 A Vtrans = -0.000095 V, Factor V/A = 0.010000 V/A, Zero Current = -0.009342 A 16:17:22 Magnet Current for Trim, Imag = -0.3492 A Vtrans = -0.029098 V, Factor V/A = 0.083333 V/A, Zero Current = 0.000000 A 16:17:40 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = -0.000129685 Vs 16:17:57 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = 0.000109942 Vs 16:17:58 Magnet Current for Main, Imag = -0.0001 A Vtrans = -0.000094 V, Factor V/A = 0.010000 V/A, Zero Current = -0.009342 A 16:17:59 Magnet Current for Trim, Imag = -0.3492 A Vtrans = -0.029098 V, Factor V/A = 0.083333 V/A, Zero Current = 0.000000 A 16:18:16 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = -0.000130750 Vs 16:18:33 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = 0.000111044 Vs 16:18:33 Magnet Current for Main, Imag = -0.0001 A Vtrans = -0.000095 V, Factor V/A = 0.010000 V/A, Zero Current = -0.009342 A 16:18:34 Magnet Current for Trim, Imag = -0.3492 A Vtrans = -0.029098 V, Factor V/A = 0.083333 V/A, Zero Current = 0.000000 A 16:18:51 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = -0.000131251 Vs 16:19:08 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = 0.000111159 Vs 16:19:09 Magnet Current for Main, Imag = -0.0000 A Vtrans = -0.000093 V, Factor V/A = 0.010000 V/A, Zero Current = -0.009342 A 16:19:10 Magnet Current for Trim, Imag = -0.3492 A Vtrans = -0.029098 V, Factor V/A = 0.083333 V/A, Zero Current = 0.000000 A 16:19:27 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = -0.000131614 Vs 16:19:44 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = 0.000110099 Vs 16:19:44 Integrated Voltage Average and RMS of Measurements X0 = 0.0000000 m, Delta X = 0.0100000 m VT = -0.000120693 +- 0.000000525 Vs 16:19:44 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0000000 m, dx = 0.0100000 m, Nturns = 10.0 VT = -0.000120693 +- 0.000000525 Vs BL = -0.00120693 +- 0.00000525 Tm 16:19:52 BL vs X Fit Coefficients: Cn sigCn n (Tm/m^n) ----- --------+-------- 0 -0.00121 0.00001 Compute Fit Residuals: X0 BL sigBL BLfit BL-BLfit (m) (Tm) (Tm) (Tm) (Tm) ----------- --------+-------- -------- -------- 0.0000000 -0.00121 0.00001 -0.00121 0.00000 16:19:52 Magnet Strength Calculation Harmonic Number = 1 Number Of X0 Positions = 1 X0 BL sigBL (m) (Tm) (Tm) -------- --------+-------- 0.0000 -0.00121 0.00001 SL = -0.0012069 +- 0.0000053 Tm^2/m^n 16:19:52 Magnet Strength Measurement Harmonic Number = 1 SL = -0.0012069 +- 0.0000053 Tm^2/m^n 16:19:52 Magnet system Main current ramping to 0.0000 A... 16:20:22 Magnet system Trim current ramping to -0.4000 A... 16:20:52 Magnetic Field Measurement Measuring the magnet strength... 16:20:52 Integrated Voltage Measurement... 16:20:53 Magnet Current for Main, Imag = -0.0001 A Vtrans = -0.000094 V, Factor V/A = 0.010000 V/A, Zero Current = -0.009342 A 16:20:54 Magnet Current for Trim, Imag = -0.3994 A Vtrans = -0.033282 V, Factor V/A = 0.083333 V/A, Zero Current = 0.000000 A 16:21:12 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = -0.000209230 Vs 16:21:29 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = 0.000186192 Vs 16:21:29 Magnet Current for Main, Imag = -0.0002 A Vtrans = -0.000095 V, Factor V/A = 0.010000 V/A, Zero Current = -0.009342 A 16:21:30 Magnet Current for Trim, Imag = -0.3994 A Vtrans = -0.033281 V, Factor V/A = 0.083333 V/A, Zero Current = 0.000000 A 16:21:47 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = -0.000209744 Vs 16:22:04 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = 0.000187661 Vs 16:22:05 Magnet Current for Main, Imag = -0.0000 A Vtrans = -0.000094 V, Factor V/A = 0.010000 V/A, Zero Current = -0.009342 A 16:22:05 Magnet Current for Trim, Imag = -0.3994 A Vtrans = -0.033281 V, Factor V/A = 0.083333 V/A, Zero Current = 0.000000 A 16:22:23 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = -0.000211059 Vs 16:22:40 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = 0.000187958 Vs 16:22:40 Magnet Current for Main, Imag = -0.0001 A Vtrans = -0.000094 V, Factor V/A = 0.010000 V/A, Zero Current = -0.009342 A 16:22:41 Magnet Current for Trim, Imag = -0.3994 A Vtrans = -0.033282 V, Factor V/A = 0.083333 V/A, Zero Current = 0.000000 A 16:22:58 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = -0.000211553 Vs 16:23:15 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = 0.000188818 Vs 16:23:15 Integrated Voltage Average and RMS of Measurements X0 = 0.0000000 m, Delta X = 0.0100000 m VT = -0.000199027 +- 0.000000924 Vs 16:23:15 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0000000 m, dx = 0.0100000 m, Nturns = 10.0 VT = -0.000199027 +- 0.000000924 Vs BL = -0.00199027 +- 0.00000924 Tm 16:23:23 BL vs X Fit Coefficients: Cn sigCn n (Tm/m^n) ----- --------+-------- 0 -0.00199 0.00001 Compute Fit Residuals: X0 BL sigBL BLfit BL-BLfit (m) (Tm) (Tm) (Tm) (Tm) ----------- --------+-------- -------- -------- 0.0000000 -0.00199 0.00001 -0.00199 0.00000 16:23:23 Magnet Strength Calculation Harmonic Number = 1 Number Of X0 Positions = 1 X0 BL sigBL (m) (Tm) (Tm) -------- --------+-------- 0.0000 -0.00199 0.00001 SL = -0.0019903 +- 0.0000092 Tm^2/m^n 16:23:23 Magnet Strength Measurement Harmonic Number = 1 SL = -0.0019903 +- 0.0000092 Tm^2/m^n 16:23:23 Magnet system Main current ramping to 0.0000 A... 16:23:53 Magnet system Trim current ramping to -0.4500 A... 16:24:24 Magnetic Field Measurement Measuring the magnet strength... 16:24:24 Integrated Voltage Measurement... 16:24:24 Magnet Current for Main, Imag = -0.0001 A Vtrans = -0.000095 V, Factor V/A = 0.010000 V/A, Zero Current = -0.009342 A 16:24:25 Magnet Current for Trim, Imag = -0.4506 A Vtrans = -0.037550 V, Factor V/A = 0.083333 V/A, Zero Current = 0.000000 A 16:24:43 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = -0.000288878 Vs 16:25:00 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = 0.000267494 Vs 16:25:00 Magnet Current for Main, Imag = -0.0001 A Vtrans = -0.000094 V, Factor V/A = 0.010000 V/A, Zero Current = -0.009342 A 16:25:01 Magnet Current for Trim, Imag = -0.4506 A Vtrans = -0.037550 V, Factor V/A = 0.083333 V/A, Zero Current = 0.000000 A 16:25:18 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = -0.000290887 Vs 16:25:35 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = 0.000267612 Vs 16:25:36 Magnet Current for Main, Imag = -0.0001 A Vtrans = -0.000095 V, Factor V/A = 0.010000 V/A, Zero Current = -0.009342 A 16:25:37 Magnet Current for Trim, Imag = -0.4506 A Vtrans = -0.037551 V, Factor V/A = 0.083333 V/A, Zero Current = 0.000000 A 16:25:54 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = -0.000291014 Vs 16:26:11 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = 0.000269280 Vs 16:26:11 Magnet Current for Main, Imag = -0.0001 A Vtrans = -0.000094 V, Factor V/A = 0.010000 V/A, Zero Current = -0.009342 A 16:26:12 Magnet Current for Trim, Imag = -0.4506 A Vtrans = -0.037550 V, Factor V/A = 0.083333 V/A, Zero Current = 0.000000 A 16:26:29 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = -0.000290924 Vs 16:26:46 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = 0.000269235 Vs 16:26:46 Integrated Voltage Average and RMS of Measurements X0 = 0.0000000 m, Delta X = 0.0100000 m VT = -0.000279416 +- 0.000000793 Vs 16:26:46 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0000000 m, dx = 0.0100000 m, Nturns = 10.0 VT = -0.000279416 +- 0.000000793 Vs BL = -0.00279416 +- 0.00000793 Tm 16:26:54 BL vs X Fit Coefficients: Cn sigCn n (Tm/m^n) ----- --------+-------- 0 -0.00279 0.00001 Compute Fit Residuals: X0 BL sigBL BLfit BL-BLfit (m) (Tm) (Tm) (Tm) (Tm) ----------- --------+-------- -------- -------- 0.0000000 -0.00279 0.00001 -0.00279 0.00000 16:26:54 Magnet Strength Calculation Harmonic Number = 1 Number Of X0 Positions = 1 X0 BL sigBL (m) (Tm) (Tm) -------- --------+-------- 0.0000 -0.00279 0.00001 SL = -0.0027942 +- 0.0000079 Tm^2/m^n 16:26:54 Magnet Strength Measurement Harmonic Number = 1 SL = -0.0027942 +- 0.0000079 Tm^2/m^n 16:26:54 Magnet system Main current ramping to 0.0000 A... 16:27:24 Magnet system Trim current ramping to -0.5000 A... 16:27:55 Magnetic Field Measurement Measuring the magnet strength... 16:27:55 Integrated Voltage Measurement... 16:27:55 Magnet Current for Main, Imag = -0.0001 A Vtrans = -0.000095 V, Factor V/A = 0.010000 V/A, Zero Current = -0.009342 A 16:27:56 Magnet Current for Trim, Imag = -0.4988 A Vtrans = -0.041563 V, Factor V/A = 0.083333 V/A, Zero Current = 0.000000 A 16:28:14 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = -0.000364859 Vs 16:28:31 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = 0.000342512 Vs 16:28:31 Magnet Current for Main, Imag = -0.0001 A Vtrans = -0.000094 V, Factor V/A = 0.010000 V/A, Zero Current = -0.009342 A 16:28:32 Magnet Current for Trim, Imag = -0.4988 A Vtrans = -0.041563 V, Factor V/A = 0.083333 V/A, Zero Current = 0.000000 A 16:28:49 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = -0.000365800 Vs 16:29:06 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = 0.000343385 Vs 16:29:07 Magnet Current for Main, Imag = -0.0001 A Vtrans = -0.000094 V, Factor V/A = 0.010000 V/A, Zero Current = -0.009342 A 16:29:08 Magnet Current for Trim, Imag = -0.4988 A Vtrans = -0.041563 V, Factor V/A = 0.083333 V/A, Zero Current = 0.000000 A 16:29:25 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = -0.000366274 Vs 16:29:42 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = 0.000343327 Vs 16:29:42 Magnet Current for Main, Imag = -0.0001 A Vtrans = -0.000095 V, Factor V/A = 0.010000 V/A, Zero Current = -0.009342 A 16:29:43 Magnet Current for Trim, Imag = -0.4988 A Vtrans = -0.041564 V, Factor V/A = 0.083333 V/A, Zero Current = 0.000000 A 16:30:00 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = -0.000367311 Vs 16:30:17 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = 0.000343374 Vs 16:30:17 Integrated Voltage Average and RMS of Measurements X0 = 0.0000000 m, Delta X = 0.0100000 m VT = -0.000354605 +- 0.000000597 Vs 16:30:17 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0000000 m, dx = 0.0100000 m, Nturns = 10.0 VT = -0.000354605 +- 0.000000597 Vs BL = -0.00354605 +- 0.00000597 Tm 16:30:25 BL vs X Fit Coefficients: Cn sigCn n (Tm/m^n) ----- --------+-------- 0 -0.00355 0.00001 Compute Fit Residuals: X0 BL sigBL BLfit BL-BLfit (m) (Tm) (Tm) (Tm) (Tm) ----------- --------+-------- -------- -------- 0.0000000 -0.00355 0.00001 -0.00355 0.00000 16:30:25 Magnet Strength Calculation Harmonic Number = 1 Number Of X0 Positions = 1 X0 BL sigBL (m) (Tm) (Tm) -------- --------+-------- 0.0000 -0.00355 0.00001 SL = -0.0035461 +- 0.0000060 Tm^2/m^n 16:30:25 Magnet Strength Measurement Harmonic Number = 1 SL = -0.0035461 +- 0.0000060 Tm^2/m^n 16:30:36 Name = None, is zeroed 16:30:36 Magnet system Main current ramping to 0.0000 A... 16:30:37 Magnet system Trim current ramping to 0.0000 A...