SLAC Magnetic Measurements Date: 06-17-2022 Time: 13:50:02 Project: Facet_II Magnet Type: Dipole Magnet Name: 1.18D3.17-BCX10481-4 Serial Number: 4 Run Number: 1 Measurement Device: 10wire Operator: sda Comment: Inhtegrated field from 0 to 185 to 0 amps in 15 amp steps. Integated field vs X at 155 amps. 2 hr warmup at 165 A Standardization for MAIN: 3 cycles from 0.0 A to 185.0 A to 0.0 A Test Currents for MAIN (A): 0.00 15.00 30.00 45.00 60.00 75.00 90.00 105.00 120.00 135.00 150.00 165.00 180.00 185.00 180.00 165.00 150.00 135.00 120.00 105.00 90.00 75.00 60.00 45.00 30.00 15.00 0.00 155.00 155.00 155.00 13:50:28 Position moved: 0.000000 m 13:50:55 Magnet Current for MAIN, Imag = 0.0052 A Vtrans = 0.000052 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 13:50:56 Magnet Current for MAIN, Imag = 0.0054 A Vtrans = 0.000054 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 13:50:56 Magnet Current for MAIN, Imag = 0.0054 A Vtrans = 0.000054 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 13:50:57 Magnet Current for MAIN, Imag = 0.0054 A Vtrans = 0.000054 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 13:50:57 Magnet Avgerage Current Measurement for MAIN : Imag = 0.0053 +- 0.0001 A 13:51:24 Initial Zero Current Measurement for MAIN, Imag_zero = 0.00534565 A 13:51:24 Ambient Measurement Device(s) Being Used, type = None 13:51:24 Magnet Voltage Measurement Device(s) Being Used, type = HP34970 13:51:25 Temperature Measurement Device(s) Being Used, type = HPTHERM_HP34970 13:51:25 XMAG Device(s) Being Used, type = None 13:51:25 Ambient Measurement Device(s) Being Used, type = HP34970 13:51:25 Magnet system MAIN current ramping to 165.0000 A... 13:51:37 Magnet Current for MAIN, Imag = 164.9726 A Vtrans = 1.649779 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 13:51:38 MAIN (V) Voltage Measurement, V = 6.500592 V 13:51:44 Temperature Measurement, Probe Number = 0, T = 25.241000 C 13:51:45 Temperature Measurement, Probe Number = 1, T = 27.027000 C 13:51:46 Temperature Measurement, Probe Number = 2, T = 27.074000 C 13:51:46 Temperature Measurement, Probe Number = 3, T = 28.482000 C 13:51:47 Temperature Measurement, Probe Number = 4, T = 28.584000 C 13:51:48 Temperature Measurement, Probe Number = 5, T = 23.499000 C 13:51:48 Sensor Voltage Measurement, Sensor Number = 0, V = 2.653993 V 13:51:48 Sensor Measurement, Sensor Number = 0, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 2.653993 (V, A, ohm, ...), Measured Value = 124.049512 (C, atm, Hz, ...) 13:51:49 Sensor Voltage Measurement, Sensor Number = 1, V = 1.153668 V 13:51:49 Sensor Measurement, Sensor Number = 1, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 1.153668 (V, A, ohm, ...), Measured Value = 11.525093 (C, atm, Hz, ...) 13:51:49 Magnet Test Parameter Measurements P0 = MAIN, Value = 164.97 P1 = MAIN (V), Value = 6.50 P2 = Yoke, Value = 25.24 P3 = Bottom Coil, Value = 27.03 P4 = Top Coil, Value = 27.07 P5 = Water In, Value = 28.48 P6 = Water Out, Value = 28.58 P7 = Ambient, Value = 23.50 P8 = Supply Pres(psi), Value = 124.05 P9 = Return Pres(psi), Value = 11.53 13:52:50 Magnet Current for MAIN, Imag = 164.9737 A Vtrans = 1.649790 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 13:52:50 MAIN (V) Voltage Measurement, V = 6.509144 V 13:52:54 Temperature Measurement, Probe Number = 0, T = 25.200000 C 13:52:54 Temperature Measurement, Probe Number = 1, T = 27.382000 C 13:52:55 Temperature Measurement, Probe Number = 2, T = 27.503000 C 13:52:55 Temperature Measurement, Probe Number = 3, T = 29.578000 C 13:52:56 Temperature Measurement, Probe Number = 4, T = 28.620000 C 13:52:57 Temperature Measurement, Probe Number = 5, T = 23.561000 C 13:52:57 Sensor Voltage Measurement, Sensor Number = 0, V = 2.657651 V 13:52:57 Sensor Measurement, Sensor Number = 0, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 2.657651 (V, A, ohm, ...), Measured Value = 124.323818 (C, atm, Hz, ...) 13:52:58 Sensor Voltage Measurement, Sensor Number = 1, V = 1.156462 V 13:52:58 Sensor Measurement, Sensor Number = 1, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 1.156462 (V, A, ohm, ...), Measured Value = 11.734680 (C, atm, Hz, ...) 13:52:58 Magnet Test Parameter Measurements P0 = MAIN, Value = 164.97 P1 = MAIN (V), Value = 6.51 P2 = Yoke, Value = 25.20 P3 = Bottom Coil, Value = 27.38 P4 = Top Coil, Value = 27.50 P5 = Water In, Value = 29.58 P6 = Water Out, Value = 28.62 P7 = Ambient, Value = 23.56 P8 = Supply Pres(psi), Value = 124.32 P9 = Return Pres(psi), Value = 11.73 13:53:59 Magnet Current for MAIN, Imag = 164.9737 A Vtrans = 1.649790 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 13:53:59 MAIN (V) Voltage Measurement, V = 6.509248 V 13:54:03 Temperature Measurement, Probe Number = 0, T = 25.212000 C 13:54:03 Temperature Measurement, Probe Number = 1, T = 27.665000 C 13:54:04 Temperature Measurement, Probe Number = 2, T = 27.755000 C 13:54:04 Temperature Measurement, Probe Number = 3, T = 30.054000 C 13:54:05 Temperature Measurement, Probe Number = 4, T = 28.615000 C 13:54:06 Temperature Measurement, Probe Number = 5, T = 23.500000 C 13:54:06 Sensor Voltage Measurement, Sensor Number = 0, V = 2.647096 V 13:54:06 Sensor Measurement, Sensor Number = 0, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 2.647096 (V, A, ohm, ...), Measured Value = 123.532193 (C, atm, Hz, ...) 13:54:07 Sensor Voltage Measurement, Sensor Number = 1, V = 1.155591 V 13:54:07 Sensor Measurement, Sensor Number = 1, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 1.155591 (V, A, ohm, ...), Measured Value = 11.669317 (C, atm, Hz, ...) 13:54:07 Magnet Test Parameter Measurements P0 = MAIN, Value = 164.97 P1 = MAIN (V), Value = 6.51 P2 = Yoke, Value = 25.21 P3 = Bottom Coil, Value = 27.66 P4 = Top Coil, Value = 27.75 P5 = Water In, Value = 30.05 P6 = Water Out, Value = 28.61 P7 = Ambient, Value = 23.50 P8 = Supply Pres(psi), Value = 123.53 P9 = Return Pres(psi), Value = 11.67 13:55:08 Magnet Current for MAIN, Imag = 164.9735 A Vtrans = 1.649789 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 13:55:08 MAIN (V) Voltage Measurement, V = 6.509550 V 13:55:12 Temperature Measurement, Probe Number = 0, T = 25.230000 C 13:55:12 Temperature Measurement, Probe Number = 1, T = 27.834000 C 13:55:13 Temperature Measurement, Probe Number = 2, T = 27.699000 C 13:55:13 Temperature Measurement, Probe Number = 3, T = 30.249000 C 13:55:14 Temperature Measurement, Probe Number = 4, T = 28.632000 C 13:55:15 Temperature Measurement, Probe Number = 5, T = 23.558000 C 13:55:15 Sensor Voltage Measurement, Sensor Number = 0, V = 2.646086 V 13:55:15 Sensor Measurement, Sensor Number = 0, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 2.646086 (V, A, ohm, ...), Measured Value = 123.456428 (C, atm, Hz, ...) 13:55:16 Sensor Voltage Measurement, Sensor Number = 1, V = 1.153438 V 13:55:16 Sensor Measurement, Sensor Number = 1, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 1.153438 (V, A, ohm, ...), Measured Value = 11.507865 (C, atm, Hz, ...) 13:55:16 Magnet Test Parameter Measurements P0 = MAIN, Value = 164.97 P1 = MAIN (V), Value = 6.51 P2 = Yoke, Value = 25.23 P3 = Bottom Coil, Value = 27.83 P4 = Top Coil, Value = 27.70 P5 = Water In, Value = 30.25 P6 = Water Out, Value = 28.63 P7 = Ambient, Value = 23.56 P8 = Supply Pres(psi), Value = 123.46 P9 = Return Pres(psi), Value = 11.51 13:56:17 Magnet Current for MAIN, Imag = 164.9732 A Vtrans = 1.649785 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 13:56:17 MAIN (V) Voltage Measurement, V = 6.509797 V 13:56:21 Temperature Measurement, Probe Number = 0, T = 25.234000 C 13:56:21 Temperature Measurement, Probe Number = 1, T = 27.980000 C 13:56:22 Temperature Measurement, Probe Number = 2, T = 27.764000 C 13:56:22 Temperature Measurement, Probe Number = 3, T = 30.336000 C 13:56:23 Temperature Measurement, Probe Number = 4, T = 28.626000 C 13:56:24 Temperature Measurement, Probe Number = 5, T = 23.612000 C 13:56:24 Sensor Voltage Measurement, Sensor Number = 0, V = 2.632519 V 13:56:24 Sensor Measurement, Sensor Number = 0, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 2.632519 (V, A, ohm, ...), Measured Value = 122.438925 (C, atm, Hz, ...) 13:56:25 Sensor Voltage Measurement, Sensor Number = 1, V = 1.152907 V 13:56:25 Sensor Measurement, Sensor Number = 1, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 1.152907 (V, A, ohm, ...), Measured Value = 11.468062 (C, atm, Hz, ...) 13:56:25 Magnet Test Parameter Measurements P0 = MAIN, Value = 164.97 P1 = MAIN (V), Value = 6.51 P2 = Yoke, Value = 25.23 P3 = Bottom Coil, Value = 27.98 P4 = Top Coil, Value = 27.76 P5 = Water In, Value = 30.34 P6 = Water Out, Value = 28.63 P7 = Ambient, Value = 23.61 P8 = Supply Pres(psi), Value = 122.44 P9 = Return Pres(psi), Value = 11.47 13:57:26 Magnet Current for MAIN, Imag = 164.9742 A Vtrans = 1.649795 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 13:57:26 MAIN (V) Voltage Measurement, V = 6.509944 V 13:57:30 Temperature Measurement, Probe Number = 0, T = 25.232000 C 13:57:30 Temperature Measurement, Probe Number = 1, T = 28.069000 C 13:57:31 Temperature Measurement, Probe Number = 2, T = 27.726000 C 13:57:31 Temperature Measurement, Probe Number = 3, T = 30.374000 C 13:57:32 Temperature Measurement, Probe Number = 4, T = 28.634000 C 13:57:33 Temperature Measurement, Probe Number = 5, T = 23.519000 C 13:57:33 Sensor Voltage Measurement, Sensor Number = 0, V = 2.651189 V 13:57:33 Sensor Measurement, Sensor Number = 0, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 2.651189 (V, A, ohm, ...), Measured Value = 123.839175 (C, atm, Hz, ...) 13:57:34 Sensor Voltage Measurement, Sensor Number = 1, V = 1.152568 V 13:57:34 Sensor Measurement, Sensor Number = 1, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 1.152568 (V, A, ohm, ...), Measured Value = 11.442585 (C, atm, Hz, ...) 13:57:34 Magnet Test Parameter Measurements P0 = MAIN, Value = 164.97 P1 = MAIN (V), Value = 6.51 P2 = Yoke, Value = 25.23 P3 = Bottom Coil, Value = 28.07 P4 = Top Coil, Value = 27.73 P5 = Water In, Value = 30.37 P6 = Water Out, Value = 28.63 P7 = Ambient, Value = 23.52 P8 = Supply Pres(psi), Value = 123.84 P9 = Return Pres(psi), Value = 11.44 13:58:35 Magnet Current for MAIN, Imag = 164.9737 A Vtrans = 1.649790 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 13:58:35 MAIN (V) Voltage Measurement, V = 6.510125 V 13:58:39 Temperature Measurement, Probe Number = 0, T = 25.233000 C 13:58:39 Temperature Measurement, Probe Number = 1, T = 28.137000 C 13:58:40 Temperature Measurement, Probe Number = 2, T = 27.787000 C 13:58:40 Temperature Measurement, Probe Number = 3, T = 30.394000 C 13:58:41 Temperature Measurement, Probe Number = 4, T = 28.649000 C 13:58:42 Temperature Measurement, Probe Number = 5, T = 23.526000 C 13:58:42 Sensor Voltage Measurement, Sensor Number = 0, V = 2.656153 V 13:58:42 Sensor Measurement, Sensor Number = 0, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 2.656153 (V, A, ohm, ...), Measured Value = 124.211438 (C, atm, Hz, ...) 13:58:43 Sensor Voltage Measurement, Sensor Number = 1, V = 1.153983 V 13:58:43 Sensor Measurement, Sensor Number = 1, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 1.153983 (V, A, ohm, ...), Measured Value = 11.548695 (C, atm, Hz, ...) 13:58:43 Magnet Test Parameter Measurements P0 = MAIN, Value = 164.97 P1 = MAIN (V), Value = 6.51 P2 = Yoke, Value = 25.23 P3 = Bottom Coil, Value = 28.14 P4 = Top Coil, Value = 27.79 P5 = Water In, Value = 30.39 P6 = Water Out, Value = 28.65 P7 = Ambient, Value = 23.53 P8 = Supply Pres(psi), Value = 124.21 P9 = Return Pres(psi), Value = 11.55 13:59:44 Magnet Current for MAIN, Imag = 164.9739 A Vtrans = 1.649793 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 13:59:44 MAIN (V) Voltage Measurement, V = 6.510182 V 13:59:48 Temperature Measurement, Probe Number = 0, T = 25.249000 C 13:59:48 Temperature Measurement, Probe Number = 1, T = 28.214000 C 13:59:49 Temperature Measurement, Probe Number = 2, T = 27.876000 C 13:59:50 Temperature Measurement, Probe Number = 3, T = 30.411000 C 13:59:50 Temperature Measurement, Probe Number = 4, T = 28.589000 C 13:59:51 Temperature Measurement, Probe Number = 5, T = 23.649000 C 13:59:51 Sensor Voltage Measurement, Sensor Number = 0, V = 2.642904 V 13:59:51 Sensor Measurement, Sensor Number = 0, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 2.642904 (V, A, ohm, ...), Measured Value = 123.217807 (C, atm, Hz, ...) 13:59:52 Sensor Voltage Measurement, Sensor Number = 1, V = 1.152448 V 13:59:52 Sensor Measurement, Sensor Number = 1, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 1.152448 (V, A, ohm, ...), Measured Value = 11.433600 (C, atm, Hz, ...) 13:59:52 Magnet Test Parameter Measurements P0 = MAIN, Value = 164.97 P1 = MAIN (V), Value = 6.51 P2 = Yoke, Value = 25.25 P3 = Bottom Coil, Value = 28.21 P4 = Top Coil, Value = 27.88 P5 = Water In, Value = 30.41 P6 = Water Out, Value = 28.59 P7 = Ambient, Value = 23.65 P8 = Supply Pres(psi), Value = 123.22 P9 = Return Pres(psi), Value = 11.43 14:00:53 Magnet Current for MAIN, Imag = 164.9738 A Vtrans = 1.649791 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 14:00:53 MAIN (V) Voltage Measurement, V = 6.510392 V 14:00:57 Temperature Measurement, Probe Number = 0, T = 25.258000 C 14:00:57 Temperature Measurement, Probe Number = 1, T = 28.267000 C 14:00:58 Temperature Measurement, Probe Number = 2, T = 27.897000 C 14:00:59 Temperature Measurement, Probe Number = 3, T = 30.425000 C 14:00:59 Temperature Measurement, Probe Number = 4, T = 28.602000 C 14:01:00 Temperature Measurement, Probe Number = 5, T = 23.611000 C 14:01:00 Sensor Voltage Measurement, Sensor Number = 0, V = 2.632012 V 14:01:00 Sensor Measurement, Sensor Number = 0, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 2.632012 (V, A, ohm, ...), Measured Value = 122.400900 (C, atm, Hz, ...) 14:01:01 Sensor Voltage Measurement, Sensor Number = 1, V = 1.152806 V 14:01:01 Sensor Measurement, Sensor Number = 1, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 1.152806 (V, A, ohm, ...), Measured Value = 11.460472 (C, atm, Hz, ...) 14:01:01 Magnet Test Parameter Measurements P0 = MAIN, Value = 164.97 P1 = MAIN (V), Value = 6.51 P2 = Yoke, Value = 25.26 P3 = Bottom Coil, Value = 28.27 P4 = Top Coil, Value = 27.90 P5 = Water In, Value = 30.43 P6 = Water Out, Value = 28.60 P7 = Ambient, Value = 23.61 P8 = Supply Pres(psi), Value = 122.40 P9 = Return Pres(psi), Value = 11.46 14:02:02 Magnet Current for MAIN, Imag = 164.9743 A Vtrans = 1.649796 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 14:02:02 MAIN (V) Voltage Measurement, V = 6.510271 V 14:02:06 Temperature Measurement, Probe Number = 0, T = 25.237000 C 14:02:06 Temperature Measurement, Probe Number = 1, T = 28.279000 C 14:02:07 Temperature Measurement, Probe Number = 2, T = 27.959000 C 14:02:08 Temperature Measurement, Probe Number = 3, T = 30.419000 C 14:02:08 Temperature Measurement, Probe Number = 4, T = 28.604000 C 14:02:09 Temperature Measurement, Probe Number = 5, T = 23.599000 C 14:02:09 Sensor Voltage Measurement, Sensor Number = 0, V = 2.659908 V 14:02:09 Sensor Measurement, Sensor Number = 0, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 2.659908 (V, A, ohm, ...), Measured Value = 124.493137 (C, atm, Hz, ...) 14:02:10 Sensor Voltage Measurement, Sensor Number = 1, V = 1.153780 V 14:02:10 Sensor Measurement, Sensor Number = 1, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 1.153780 (V, A, ohm, ...), Measured Value = 11.533522 (C, atm, Hz, ...) 14:02:10 Magnet Test Parameter Measurements P0 = MAIN, Value = 164.97 P1 = MAIN (V), Value = 6.51 P2 = Yoke, Value = 25.24 P3 = Bottom Coil, Value = 28.28 P4 = Top Coil, Value = 27.96 P5 = Water In, Value = 30.42 P6 = Water Out, Value = 28.60 P7 = Ambient, Value = 23.60 P8 = Supply Pres(psi), Value = 124.49 P9 = Return Pres(psi), Value = 11.53 14:03:11 Magnet Current for MAIN, Imag = 164.9737 A Vtrans = 1.649790 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 14:03:11 MAIN (V) Voltage Measurement, V = 6.510451 V 14:03:15 Temperature Measurement, Probe Number = 0, T = 25.250000 C 14:03:15 Temperature Measurement, Probe Number = 1, T = 28.284000 C 14:03:16 Temperature Measurement, Probe Number = 2, T = 27.905000 C 14:03:17 Temperature Measurement, Probe Number = 3, T = 30.426000 C 14:03:17 Temperature Measurement, Probe Number = 4, T = 28.603000 C 14:03:18 Temperature Measurement, Probe Number = 5, T = 23.562000 C 14:03:18 Sensor Voltage Measurement, Sensor Number = 0, V = 2.662241 V 14:03:18 Sensor Measurement, Sensor Number = 0, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 2.662241 (V, A, ohm, ...), Measured Value = 124.668075 (C, atm, Hz, ...) 14:03:19 Sensor Voltage Measurement, Sensor Number = 1, V = 1.155946 V 14:03:19 Sensor Measurement, Sensor Number = 1, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 1.155946 (V, A, ohm, ...), Measured Value = 11.695913 (C, atm, Hz, ...) 14:03:19 Magnet Test Parameter Measurements P0 = MAIN, Value = 164.97 P1 = MAIN (V), Value = 6.51 P2 = Yoke, Value = 25.25 P3 = Bottom Coil, Value = 28.28 P4 = Top Coil, Value = 27.91 P5 = Water In, Value = 30.43 P6 = Water Out, Value = 28.60 P7 = Ambient, Value = 23.56 P8 = Supply Pres(psi), Value = 124.67 P9 = Return Pres(psi), Value = 11.70 14:04:20 Magnet Current for MAIN, Imag = 164.9743 A Vtrans = 1.649796 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 14:04:20 MAIN (V) Voltage Measurement, V = 6.510487 V 14:04:24 Temperature Measurement, Probe Number = 0, T = 25.274000 C 14:04:24 Temperature Measurement, Probe Number = 1, T = 28.277000 C 14:04:25 Temperature Measurement, Probe Number = 2, T = 27.923000 C 14:04:26 Temperature Measurement, Probe Number = 3, T = 30.400000 C 14:04:26 Temperature Measurement, Probe Number = 4, T = 28.636000 C 14:04:27 Temperature Measurement, Probe Number = 5, T = 23.587000 C 14:04:27 Sensor Voltage Measurement, Sensor Number = 0, V = 2.664316 V 14:04:27 Sensor Measurement, Sensor Number = 0, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 2.664316 (V, A, ohm, ...), Measured Value = 124.823722 (C, atm, Hz, ...) 14:04:28 Sensor Voltage Measurement, Sensor Number = 1, V = 1.154943 V 14:04:28 Sensor Measurement, Sensor Number = 1, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 1.154943 (V, A, ohm, ...), Measured Value = 11.620710 (C, atm, Hz, ...) 14:04:28 Magnet Test Parameter Measurements P0 = MAIN, Value = 164.97 P1 = MAIN (V), Value = 6.51 P2 = Yoke, Value = 25.27 P3 = Bottom Coil, Value = 28.28 P4 = Top Coil, Value = 27.92 P5 = Water In, Value = 30.40 P6 = Water Out, Value = 28.64 P7 = Ambient, Value = 23.59 P8 = Supply Pres(psi), Value = 124.82 P9 = Return Pres(psi), Value = 11.62 14:05:29 Magnet Current for MAIN, Imag = 164.9740 A Vtrans = 1.649794 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 14:05:29 MAIN (V) Voltage Measurement, V = 6.510418 V 14:05:33 Temperature Measurement, Probe Number = 0, T = 25.287000 C 14:05:33 Temperature Measurement, Probe Number = 1, T = 28.296000 C 14:05:34 Temperature Measurement, Probe Number = 2, T = 27.862000 C 14:05:35 Temperature Measurement, Probe Number = 3, T = 30.423000 C 14:05:35 Temperature Measurement, Probe Number = 4, T = 28.659000 C 14:05:36 Temperature Measurement, Probe Number = 5, T = 23.590000 C 14:05:36 Sensor Voltage Measurement, Sensor Number = 0, V = 2.647610 V 14:05:36 Sensor Measurement, Sensor Number = 0, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 2.647610 (V, A, ohm, ...), Measured Value = 123.570780 (C, atm, Hz, ...) 14:05:37 Sensor Voltage Measurement, Sensor Number = 1, V = 1.154361 V 14:05:37 Sensor Measurement, Sensor Number = 1, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 1.154361 (V, A, ohm, ...), Measured Value = 11.577067 (C, atm, Hz, ...) 14:05:37 Magnet Test Parameter Measurements P0 = MAIN, Value = 164.97 P1 = MAIN (V), Value = 6.51 P2 = Yoke, Value = 25.29 P3 = Bottom Coil, Value = 28.30 P4 = Top Coil, Value = 27.86 P5 = Water In, Value = 30.42 P6 = Water Out, Value = 28.66 P7 = Ambient, Value = 23.59 P8 = Supply Pres(psi), Value = 123.57 P9 = Return Pres(psi), Value = 11.58 14:06:38 Magnet Current for MAIN, Imag = 164.9740 A Vtrans = 1.649794 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 14:06:38 MAIN (V) Voltage Measurement, V = 6.510568 V 14:06:42 Temperature Measurement, Probe Number = 0, T = 25.300000 C 14:06:42 Temperature Measurement, Probe Number = 1, T = 28.304000 C 14:06:43 Temperature Measurement, Probe Number = 2, T = 27.841000 C 14:06:44 Temperature Measurement, Probe Number = 3, T = 30.448000 C 14:06:44 Temperature Measurement, Probe Number = 4, T = 28.639000 C 14:06:45 Temperature Measurement, Probe Number = 5, T = 23.588000 C 14:06:45 Sensor Voltage Measurement, Sensor Number = 0, V = 2.660013 V 14:06:45 Sensor Measurement, Sensor Number = 0, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 2.660013 (V, A, ohm, ...), Measured Value = 124.501005 (C, atm, Hz, ...) 14:06:46 Sensor Voltage Measurement, Sensor Number = 1, V = 1.154378 V 14:06:46 Sensor Measurement, Sensor Number = 1, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 1.154378 (V, A, ohm, ...), Measured Value = 11.578380 (C, atm, Hz, ...) 14:06:46 Magnet Test Parameter Measurements P0 = MAIN, Value = 164.97 P1 = MAIN (V), Value = 6.51 P2 = Yoke, Value = 25.30 P3 = Bottom Coil, Value = 28.30 P4 = Top Coil, Value = 27.84 P5 = Water In, Value = 30.45 P6 = Water Out, Value = 28.64 P7 = Ambient, Value = 23.59 P8 = Supply Pres(psi), Value = 124.50 P9 = Return Pres(psi), Value = 11.58 14:07:47 Magnet Current for MAIN, Imag = 164.9737 A Vtrans = 1.649790 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 14:07:47 MAIN (V) Voltage Measurement, V = 6.510740 V 14:07:51 Temperature Measurement, Probe Number = 0, T = 25.296000 C 14:07:51 Temperature Measurement, Probe Number = 1, T = 28.327000 C 14:07:52 Temperature Measurement, Probe Number = 2, T = 27.949000 C 14:07:53 Temperature Measurement, Probe Number = 3, T = 30.462000 C 14:07:53 Temperature Measurement, Probe Number = 4, T = 28.637000 C 14:07:54 Temperature Measurement, Probe Number = 5, T = 23.567000 C 14:07:54 Sensor Voltage Measurement, Sensor Number = 0, V = 2.646274 V 14:07:54 Sensor Measurement, Sensor Number = 0, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 2.646274 (V, A, ohm, ...), Measured Value = 123.470573 (C, atm, Hz, ...) 14:07:55 Sensor Voltage Measurement, Sensor Number = 1, V = 1.155730 V 14:07:55 Sensor Measurement, Sensor Number = 1, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 1.155730 (V, A, ohm, ...), Measured Value = 11.679713 (C, atm, Hz, ...) 14:07:55 Magnet Test Parameter Measurements P0 = MAIN, Value = 164.97 P1 = MAIN (V), Value = 6.51 P2 = Yoke, Value = 25.30 P3 = Bottom Coil, Value = 28.33 P4 = Top Coil, Value = 27.95 P5 = Water In, Value = 30.46 P6 = Water Out, Value = 28.64 P7 = Ambient, Value = 23.57 P8 = Supply Pres(psi), Value = 123.47 P9 = Return Pres(psi), Value = 11.68 14:08:56 Magnet Current for MAIN, Imag = 164.9734 A Vtrans = 1.649788 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 14:08:56 MAIN (V) Voltage Measurement, V = 6.510567 V 14:09:00 Temperature Measurement, Probe Number = 0, T = 25.324000 C 14:09:00 Temperature Measurement, Probe Number = 1, T = 28.339000 C 14:09:01 Temperature Measurement, Probe Number = 2, T = 28.196000 C 14:09:02 Temperature Measurement, Probe Number = 3, T = 30.468000 C 14:09:02 Temperature Measurement, Probe Number = 4, T = 28.609000 C 14:09:03 Temperature Measurement, Probe Number = 5, T = 23.612000 C 14:09:03 Sensor Voltage Measurement, Sensor Number = 0, V = 2.652862 V 14:09:03 Sensor Measurement, Sensor Number = 0, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 2.652862 (V, A, ohm, ...), Measured Value = 123.964665 (C, atm, Hz, ...) 14:09:04 Sensor Voltage Measurement, Sensor Number = 1, V = 1.157476 V 14:09:04 Sensor Measurement, Sensor Number = 1, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 1.157476 (V, A, ohm, ...), Measured Value = 11.810722 (C, atm, Hz, ...) 14:09:04 Magnet Test Parameter Measurements P0 = MAIN, Value = 164.97 P1 = MAIN (V), Value = 6.51 P2 = Yoke, Value = 25.32 P3 = Bottom Coil, Value = 28.34 P4 = Top Coil, Value = 28.20 P5 = Water In, Value = 30.47 P6 = Water Out, Value = 28.61 P7 = Ambient, Value = 23.61 P8 = Supply Pres(psi), Value = 123.96 P9 = Return Pres(psi), Value = 11.81 14:10:05 Magnet Current for MAIN, Imag = 164.9737 A Vtrans = 1.649790 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 14:10:05 MAIN (V) Voltage Measurement, V = 6.510597 V 14:10:09 Temperature Measurement, Probe Number = 0, T = 25.325000 C 14:10:09 Temperature Measurement, Probe Number = 1, T = 28.308000 C 14:10:10 Temperature Measurement, Probe Number = 2, T = 28.298000 C 14:10:11 Temperature Measurement, Probe Number = 3, T = 30.445000 C 14:10:11 Temperature Measurement, Probe Number = 4, T = 28.589000 C 14:10:12 Temperature Measurement, Probe Number = 5, T = 23.544000 C 14:10:12 Sensor Voltage Measurement, Sensor Number = 0, V = 2.650151 V 14:10:12 Sensor Measurement, Sensor Number = 0, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 2.650151 (V, A, ohm, ...), Measured Value = 123.761355 (C, atm, Hz, ...) 14:10:13 Sensor Voltage Measurement, Sensor Number = 1, V = 1.154502 V 14:10:13 Sensor Measurement, Sensor Number = 1, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 1.154502 (V, A, ohm, ...), Measured Value = 11.587650 (C, atm, Hz, ...) 14:10:13 Magnet Test Parameter Measurements P0 = MAIN, Value = 164.97 P1 = MAIN (V), Value = 6.51 P2 = Yoke, Value = 25.32 P3 = Bottom Coil, Value = 28.31 P4 = Top Coil, Value = 28.30 P5 = Water In, Value = 30.45 P6 = Water Out, Value = 28.59 P7 = Ambient, Value = 23.54 P8 = Supply Pres(psi), Value = 123.76 P9 = Return Pres(psi), Value = 11.59 14:11:14 Magnet Current for MAIN, Imag = 164.9742 A Vtrans = 1.649795 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 14:11:14 MAIN (V) Voltage Measurement, V = 6.510649 V 14:11:18 Temperature Measurement, Probe Number = 0, T = 25.329000 C 14:11:18 Temperature Measurement, Probe Number = 1, T = 28.296000 C 14:11:19 Temperature Measurement, Probe Number = 2, T = 28.319000 C 14:11:20 Temperature Measurement, Probe Number = 3, T = 30.416000 C 14:11:20 Temperature Measurement, Probe Number = 4, T = 28.624000 C 14:11:21 Temperature Measurement, Probe Number = 5, T = 23.604000 C 14:11:21 Sensor Voltage Measurement, Sensor Number = 0, V = 2.664291 V 14:11:21 Sensor Measurement, Sensor Number = 0, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 2.664291 (V, A, ohm, ...), Measured Value = 124.821847 (C, atm, Hz, ...) 14:11:22 Sensor Voltage Measurement, Sensor Number = 1, V = 1.157032 V 14:11:22 Sensor Measurement, Sensor Number = 1, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 1.157032 (V, A, ohm, ...), Measured Value = 11.777385 (C, atm, Hz, ...) 14:11:22 Magnet Test Parameter Measurements P0 = MAIN, Value = 164.97 P1 = MAIN (V), Value = 6.51 P2 = Yoke, Value = 25.33 P3 = Bottom Coil, Value = 28.30 P4 = Top Coil, Value = 28.32 P5 = Water In, Value = 30.42 P6 = Water Out, Value = 28.62 P7 = Ambient, Value = 23.60 P8 = Supply Pres(psi), Value = 124.82 P9 = Return Pres(psi), Value = 11.78 14:12:23 Magnet Current for MAIN, Imag = 164.9744 A Vtrans = 1.649798 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 14:12:23 MAIN (V) Voltage Measurement, V = 6.510784 V 14:12:27 Temperature Measurement, Probe Number = 0, T = 25.340000 C 14:12:27 Temperature Measurement, Probe Number = 1, T = 28.270000 C 14:12:28 Temperature Measurement, Probe Number = 2, T = 28.371000 C 14:12:29 Temperature Measurement, Probe Number = 3, T = 30.444000 C 14:12:29 Temperature Measurement, Probe Number = 4, T = 28.572000 C 14:12:30 Temperature Measurement, Probe Number = 5, T = 23.549000 C 14:12:30 Sensor Voltage Measurement, Sensor Number = 0, V = 2.662830 V 14:12:30 Sensor Measurement, Sensor Number = 0, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 2.662830 (V, A, ohm, ...), Measured Value = 124.712280 (C, atm, Hz, ...) 14:12:31 Sensor Voltage Measurement, Sensor Number = 1, V = 1.153404 V 14:12:31 Sensor Measurement, Sensor Number = 1, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 1.153404 (V, A, ohm, ...), Measured Value = 11.505330 (C, atm, Hz, ...) 14:12:31 Magnet Test Parameter Measurements P0 = MAIN, Value = 164.97 P1 = MAIN (V), Value = 6.51 P2 = Yoke, Value = 25.34 P3 = Bottom Coil, Value = 28.27 P4 = Top Coil, Value = 28.37 P5 = Water In, Value = 30.44 P6 = Water Out, Value = 28.57 P7 = Ambient, Value = 23.55 P8 = Supply Pres(psi), Value = 124.71 P9 = Return Pres(psi), Value = 11.51 14:13:32 Magnet Current for MAIN, Imag = 164.9744 A Vtrans = 1.649798 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 14:13:32 MAIN (V) Voltage Measurement, V = 6.510543 V 14:13:36 Temperature Measurement, Probe Number = 0, T = 25.308000 C 14:13:36 Temperature Measurement, Probe Number = 1, T = 28.172000 C 14:13:37 Temperature Measurement, Probe Number = 2, T = 28.442000 C 14:13:38 Temperature Measurement, Probe Number = 3, T = 30.442000 C 14:13:38 Temperature Measurement, Probe Number = 4, T = 28.545000 C 14:13:39 Temperature Measurement, Probe Number = 5, T = 23.539000 C 14:13:39 Sensor Voltage Measurement, Sensor Number = 0, V = 2.646740 V 14:13:39 Sensor Measurement, Sensor Number = 0, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 2.646740 (V, A, ohm, ...), Measured Value = 123.505500 (C, atm, Hz, ...) 14:13:40 Sensor Voltage Measurement, Sensor Number = 1, V = 1.154107 V 14:13:40 Sensor Measurement, Sensor Number = 1, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 1.154107 (V, A, ohm, ...), Measured Value = 11.558055 (C, atm, Hz, ...) 14:13:40 Magnet Test Parameter Measurements P0 = MAIN, Value = 164.97 P1 = MAIN (V), Value = 6.51 P2 = Yoke, Value = 25.31 P3 = Bottom Coil, Value = 28.17 P4 = Top Coil, Value = 28.44 P5 = Water In, Value = 30.44 P6 = Water Out, Value = 28.55 P7 = Ambient, Value = 23.54 P8 = Supply Pres(psi), Value = 123.51 P9 = Return Pres(psi), Value = 11.56 14:14:41 Magnet Current for MAIN, Imag = 164.9738 A Vtrans = 1.649791 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 14:14:41 MAIN (V) Voltage Measurement, V = 6.510444 V 14:14:45 Temperature Measurement, Probe Number = 0, T = 25.307000 C 14:14:45 Temperature Measurement, Probe Number = 1, T = 28.042000 C 14:14:46 Temperature Measurement, Probe Number = 2, T = 28.467000 C 14:14:47 Temperature Measurement, Probe Number = 3, T = 30.423000 C 14:14:47 Temperature Measurement, Probe Number = 4, T = 28.538000 C 14:14:48 Temperature Measurement, Probe Number = 5, T = 23.457000 C 14:14:48 Sensor Voltage Measurement, Sensor Number = 0, V = 2.657280 V 14:14:48 Sensor Measurement, Sensor Number = 0, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 2.657280 (V, A, ohm, ...), Measured Value = 124.296000 (C, atm, Hz, ...) 14:14:49 Sensor Voltage Measurement, Sensor Number = 1, V = 1.156136 V 14:14:49 Sensor Measurement, Sensor Number = 1, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 1.156136 (V, A, ohm, ...), Measured Value = 11.710237 (C, atm, Hz, ...) 14:14:49 Magnet Test Parameter Measurements P0 = MAIN, Value = 164.97 P1 = MAIN (V), Value = 6.51 P2 = Yoke, Value = 25.31 P3 = Bottom Coil, Value = 28.04 P4 = Top Coil, Value = 28.47 P5 = Water In, Value = 30.42 P6 = Water Out, Value = 28.54 P7 = Ambient, Value = 23.46 P8 = Supply Pres(psi), Value = 124.30 P9 = Return Pres(psi), Value = 11.71 14:15:50 Magnet Current for MAIN, Imag = 164.9739 A Vtrans = 1.649793 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 14:15:50 MAIN (V) Voltage Measurement, V = 6.510594 V 14:15:54 Temperature Measurement, Probe Number = 0, T = 25.316000 C 14:15:54 Temperature Measurement, Probe Number = 1, T = 27.748000 C 14:15:55 Temperature Measurement, Probe Number = 2, T = 28.407000 C 14:15:56 Temperature Measurement, Probe Number = 3, T = 30.438000 C 14:15:56 Temperature Measurement, Probe Number = 4, T = 28.541000 C 14:15:57 Temperature Measurement, Probe Number = 5, T = 23.554000 C 14:15:57 Sensor Voltage Measurement, Sensor Number = 0, V = 2.659395 V 14:15:57 Sensor Measurement, Sensor Number = 0, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 2.659395 (V, A, ohm, ...), Measured Value = 124.454647 (C, atm, Hz, ...) 14:15:58 Sensor Voltage Measurement, Sensor Number = 1, V = 1.154768 V 14:15:58 Sensor Measurement, Sensor Number = 1, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 1.154768 (V, A, ohm, ...), Measured Value = 11.607600 (C, atm, Hz, ...) 14:15:58 Magnet Test Parameter Measurements P0 = MAIN, Value = 164.97 P1 = MAIN (V), Value = 6.51 P2 = Yoke, Value = 25.32 P3 = Bottom Coil, Value = 27.75 P4 = Top Coil, Value = 28.41 P5 = Water In, Value = 30.44 P6 = Water Out, Value = 28.54 P7 = Ambient, Value = 23.55 P8 = Supply Pres(psi), Value = 124.45 P9 = Return Pres(psi), Value = 11.61 14:16:59 Magnet Current for MAIN, Imag = 164.9742 A Vtrans = 1.649795 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 14:16:59 MAIN (V) Voltage Measurement, V = 6.510386 V 14:17:03 Temperature Measurement, Probe Number = 0, T = 25.329000 C 14:17:03 Temperature Measurement, Probe Number = 1, T = 27.534000 C 14:17:04 Temperature Measurement, Probe Number = 2, T = 28.432000 C 14:17:05 Temperature Measurement, Probe Number = 3, T = 30.433000 C 14:17:05 Temperature Measurement, Probe Number = 4, T = 28.555000 C 14:17:06 Temperature Measurement, Probe Number = 5, T = 23.589000 C 14:17:06 Sensor Voltage Measurement, Sensor Number = 0, V = 2.647635 V 14:17:06 Sensor Measurement, Sensor Number = 0, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 2.647635 (V, A, ohm, ...), Measured Value = 123.572648 (C, atm, Hz, ...) 14:17:07 Sensor Voltage Measurement, Sensor Number = 1, V = 1.153536 V 14:17:07 Sensor Measurement, Sensor Number = 1, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 1.153536 (V, A, ohm, ...), Measured Value = 11.515170 (C, atm, Hz, ...) 14:17:07 Magnet Test Parameter Measurements P0 = MAIN, Value = 164.97 P1 = MAIN (V), Value = 6.51 P2 = Yoke, Value = 25.33 P3 = Bottom Coil, Value = 27.53 P4 = Top Coil, Value = 28.43 P5 = Water In, Value = 30.43 P6 = Water Out, Value = 28.55 P7 = Ambient, Value = 23.59 P8 = Supply Pres(psi), Value = 123.57 P9 = Return Pres(psi), Value = 11.52 14:18:08 Magnet Current for MAIN, Imag = 164.9742 A Vtrans = 1.649795 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 14:18:08 MAIN (V) Voltage Measurement, V = 6.510298 V 14:18:12 Temperature Measurement, Probe Number = 0, T = 25.340000 C 14:18:12 Temperature Measurement, Probe Number = 1, T = 27.349000 C 14:18:13 Temperature Measurement, Probe Number = 2, T = 28.325000 C 14:18:14 Temperature Measurement, Probe Number = 3, T = 30.412000 C 14:18:14 Temperature Measurement, Probe Number = 4, T = 28.604000 C 14:18:15 Temperature Measurement, Probe Number = 5, T = 23.581000 C 14:18:15 Sensor Voltage Measurement, Sensor Number = 0, V = 2.651835 V 14:18:15 Sensor Measurement, Sensor Number = 0, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 2.651835 (V, A, ohm, ...), Measured Value = 123.887595 (C, atm, Hz, ...) 14:18:16 Sensor Voltage Measurement, Sensor Number = 1, V = 1.153833 V 14:18:16 Sensor Measurement, Sensor Number = 1, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 1.153833 (V, A, ohm, ...), Measured Value = 11.537452 (C, atm, Hz, ...) 14:18:16 Magnet Test Parameter Measurements P0 = MAIN, Value = 164.97 P1 = MAIN (V), Value = 6.51 P2 = Yoke, Value = 25.34 P3 = Bottom Coil, Value = 27.35 P4 = Top Coil, Value = 28.32 P5 = Water In, Value = 30.41 P6 = Water Out, Value = 28.60 P7 = Ambient, Value = 23.58 P8 = Supply Pres(psi), Value = 123.89 P9 = Return Pres(psi), Value = 11.54 14:19:17 Magnet Current for MAIN, Imag = 164.9740 A Vtrans = 1.649794 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 14:19:17 MAIN (V) Voltage Measurement, V = 6.510342 V 14:19:21 Temperature Measurement, Probe Number = 0, T = 25.344000 C 14:19:21 Temperature Measurement, Probe Number = 1, T = 27.289000 C 14:19:22 Temperature Measurement, Probe Number = 2, T = 28.213000 C 14:19:23 Temperature Measurement, Probe Number = 3, T = 30.397000 C 14:19:23 Temperature Measurement, Probe Number = 4, T = 28.643000 C 14:19:24 Temperature Measurement, Probe Number = 5, T = 23.609000 C 14:19:24 Sensor Voltage Measurement, Sensor Number = 0, V = 2.654725 V 14:19:24 Sensor Measurement, Sensor Number = 0, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 2.654725 (V, A, ohm, ...), Measured Value = 124.104397 (C, atm, Hz, ...) 14:19:25 Sensor Voltage Measurement, Sensor Number = 1, V = 1.156012 V 14:19:25 Sensor Measurement, Sensor Number = 1, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 1.156012 (V, A, ohm, ...), Measured Value = 11.700878 (C, atm, Hz, ...) 14:19:25 Magnet Test Parameter Measurements P0 = MAIN, Value = 164.97 P1 = MAIN (V), Value = 6.51 P2 = Yoke, Value = 25.34 P3 = Bottom Coil, Value = 27.29 P4 = Top Coil, Value = 28.21 P5 = Water In, Value = 30.40 P6 = Water Out, Value = 28.64 P7 = Ambient, Value = 23.61 P8 = Supply Pres(psi), Value = 124.10 P9 = Return Pres(psi), Value = 11.70 14:20:26 Magnet Current for MAIN, Imag = 164.9739 A Vtrans = 1.649793 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 14:20:26 MAIN (V) Voltage Measurement, V = 6.510562 V 14:20:30 Temperature Measurement, Probe Number = 0, T = 25.347000 C 14:20:30 Temperature Measurement, Probe Number = 1, T = 27.197000 C 14:20:31 Temperature Measurement, Probe Number = 2, T = 28.320000 C 14:20:32 Temperature Measurement, Probe Number = 3, T = 30.418000 C 14:20:32 Temperature Measurement, Probe Number = 4, T = 28.636000 C 14:20:33 Temperature Measurement, Probe Number = 5, T = 23.623000 C 14:20:33 Sensor Voltage Measurement, Sensor Number = 0, V = 2.644616 V 14:20:33 Sensor Measurement, Sensor Number = 0, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 2.644616 (V, A, ohm, ...), Measured Value = 123.346207 (C, atm, Hz, ...) 14:20:34 Sensor Voltage Measurement, Sensor Number = 1, V = 1.156116 V 14:20:34 Sensor Measurement, Sensor Number = 1, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 1.156116 (V, A, ohm, ...), Measured Value = 11.708737 (C, atm, Hz, ...) 14:20:34 Magnet Test Parameter Measurements P0 = MAIN, Value = 164.97 P1 = MAIN (V), Value = 6.51 P2 = Yoke, Value = 25.35 P3 = Bottom Coil, Value = 27.20 P4 = Top Coil, Value = 28.32 P5 = Water In, Value = 30.42 P6 = Water Out, Value = 28.64 P7 = Ambient, Value = 23.62 P8 = Supply Pres(psi), Value = 123.35 P9 = Return Pres(psi), Value = 11.71 14:21:35 Magnet Current for MAIN, Imag = 164.9743 A Vtrans = 1.649796 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 14:21:35 MAIN (V) Voltage Measurement, V = 6.510569 V 14:21:39 Temperature Measurement, Probe Number = 0, T = 25.348000 C 14:21:39 Temperature Measurement, Probe Number = 1, T = 27.143000 C 14:21:40 Temperature Measurement, Probe Number = 2, T = 28.300000 C 14:21:41 Temperature Measurement, Probe Number = 3, T = 30.452000 C 14:21:41 Temperature Measurement, Probe Number = 4, T = 28.636000 C 14:21:42 Temperature Measurement, Probe Number = 5, T = 23.687000 C 14:21:42 Sensor Voltage Measurement, Sensor Number = 0, V = 2.652380 V 14:21:42 Sensor Measurement, Sensor Number = 0, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 2.652380 (V, A, ohm, ...), Measured Value = 123.928522 (C, atm, Hz, ...) 14:21:43 Sensor Voltage Measurement, Sensor Number = 1, V = 1.155019 V 14:21:43 Sensor Measurement, Sensor Number = 1, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 1.155019 (V, A, ohm, ...), Measured Value = 11.626425 (C, atm, Hz, ...) 14:21:43 Magnet Test Parameter Measurements P0 = MAIN, Value = 164.97 P1 = MAIN (V), Value = 6.51 P2 = Yoke, Value = 25.35 P3 = Bottom Coil, Value = 27.14 P4 = Top Coil, Value = 28.30 P5 = Water In, Value = 30.45 P6 = Water Out, Value = 28.64 P7 = Ambient, Value = 23.69 P8 = Supply Pres(psi), Value = 123.93 P9 = Return Pres(psi), Value = 11.63 14:22:44 Magnet Current for MAIN, Imag = 164.9740 A Vtrans = 1.649794 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 14:22:44 MAIN (V) Voltage Measurement, V = 6.510649 V 14:22:48 Temperature Measurement, Probe Number = 0, T = 25.359000 C 14:22:49 Temperature Measurement, Probe Number = 1, T = 27.116000 C 14:22:49 Temperature Measurement, Probe Number = 2, T = 28.270000 C 14:22:50 Temperature Measurement, Probe Number = 3, T = 30.464000 C 14:22:50 Temperature Measurement, Probe Number = 4, T = 28.663000 C 14:22:51 Temperature Measurement, Probe Number = 5, T = 23.585000 C 14:22:51 Sensor Voltage Measurement, Sensor Number = 0, V = 2.665269 V 14:22:51 Sensor Measurement, Sensor Number = 0, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 2.665269 (V, A, ohm, ...), Measured Value = 124.895175 (C, atm, Hz, ...) 14:22:52 Sensor Voltage Measurement, Sensor Number = 1, V = 1.156120 V 14:22:52 Sensor Measurement, Sensor Number = 1, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 1.156120 (V, A, ohm, ...), Measured Value = 11.709022 (C, atm, Hz, ...) 14:22:52 Magnet Test Parameter Measurements P0 = MAIN, Value = 164.97 P1 = MAIN (V), Value = 6.51 P2 = Yoke, Value = 25.36 P3 = Bottom Coil, Value = 27.12 P4 = Top Coil, Value = 28.27 P5 = Water In, Value = 30.46 P6 = Water Out, Value = 28.66 P7 = Ambient, Value = 23.59 P8 = Supply Pres(psi), Value = 124.90 P9 = Return Pres(psi), Value = 11.71 14:23:53 Magnet Current for MAIN, Imag = 164.9747 A Vtrans = 1.649800 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 14:23:53 MAIN (V) Voltage Measurement, V = 6.510547 V 14:23:57 Temperature Measurement, Probe Number = 0, T = 25.383000 C 14:23:58 Temperature Measurement, Probe Number = 1, T = 27.109000 C 14:23:58 Temperature Measurement, Probe Number = 2, T = 28.088000 C 14:23:59 Temperature Measurement, Probe Number = 3, T = 30.459000 C 14:23:59 Temperature Measurement, Probe Number = 4, T = 28.663000 C 14:24:00 Temperature Measurement, Probe Number = 5, T = 23.649000 C 14:24:00 Sensor Voltage Measurement, Sensor Number = 0, V = 2.648106 V 14:24:00 Sensor Measurement, Sensor Number = 0, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 2.648106 (V, A, ohm, ...), Measured Value = 123.607958 (C, atm, Hz, ...) 14:24:01 Sensor Voltage Measurement, Sensor Number = 1, V = 1.154838 V 14:24:01 Sensor Measurement, Sensor Number = 1, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 1.154838 (V, A, ohm, ...), Measured Value = 11.612842 (C, atm, Hz, ...) 14:24:01 Magnet Test Parameter Measurements P0 = MAIN, Value = 164.97 P1 = MAIN (V), Value = 6.51 P2 = Yoke, Value = 25.38 P3 = Bottom Coil, Value = 27.11 P4 = Top Coil, Value = 28.09 P5 = Water In, Value = 30.46 P6 = Water Out, Value = 28.66 P7 = Ambient, Value = 23.65 P8 = Supply Pres(psi), Value = 123.61 P9 = Return Pres(psi), Value = 11.61 14:25:02 Magnet Current for MAIN, Imag = 164.9747 A Vtrans = 1.649800 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 14:25:02 MAIN (V) Voltage Measurement, V = 6.510513 V 14:25:06 Temperature Measurement, Probe Number = 0, T = 25.363000 C 14:25:07 Temperature Measurement, Probe Number = 1, T = 27.127000 C 14:25:07 Temperature Measurement, Probe Number = 2, T = 27.893000 C 14:25:08 Temperature Measurement, Probe Number = 3, T = 30.466000 C 14:25:08 Temperature Measurement, Probe Number = 4, T = 28.635000 C 14:25:09 Temperature Measurement, Probe Number = 5, T = 23.602000 C 14:25:09 Sensor Voltage Measurement, Sensor Number = 0, V = 2.654399 V 14:25:09 Sensor Measurement, Sensor Number = 0, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 2.654399 (V, A, ohm, ...), Measured Value = 124.079955 (C, atm, Hz, ...) 14:25:10 Sensor Voltage Measurement, Sensor Number = 1, V = 1.155426 V 14:25:10 Sensor Measurement, Sensor Number = 1, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 1.155426 (V, A, ohm, ...), Measured Value = 11.656950 (C, atm, Hz, ...) 14:25:10 Magnet Test Parameter Measurements P0 = MAIN, Value = 164.97 P1 = MAIN (V), Value = 6.51 P2 = Yoke, Value = 25.36 P3 = Bottom Coil, Value = 27.13 P4 = Top Coil, Value = 27.89 P5 = Water In, Value = 30.47 P6 = Water Out, Value = 28.64 P7 = Ambient, Value = 23.60 P8 = Supply Pres(psi), Value = 124.08 P9 = Return Pres(psi), Value = 11.66 14:26:11 Magnet Current for MAIN, Imag = 164.9743 A Vtrans = 1.649796 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 14:26:11 MAIN (V) Voltage Measurement, V = 6.510168 V 14:26:15 Temperature Measurement, Probe Number = 0, T = 25.374000 C 14:26:16 Temperature Measurement, Probe Number = 1, T = 26.954000 C 14:26:16 Temperature Measurement, Probe Number = 2, T = 27.916000 C 14:26:17 Temperature Measurement, Probe Number = 3, T = 30.452000 C 14:26:17 Temperature Measurement, Probe Number = 4, T = 28.595000 C 14:26:18 Temperature Measurement, Probe Number = 5, T = 23.702000 C 14:26:18 Sensor Voltage Measurement, Sensor Number = 0, V = 2.655804 V 14:26:19 Sensor Measurement, Sensor Number = 0, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 2.655804 (V, A, ohm, ...), Measured Value = 124.185308 (C, atm, Hz, ...) 14:26:19 Sensor Voltage Measurement, Sensor Number = 1, V = 1.154322 V 14:26:19 Sensor Measurement, Sensor Number = 1, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 1.154322 (V, A, ohm, ...), Measured Value = 11.574165 (C, atm, Hz, ...) 14:26:19 Magnet Test Parameter Measurements P0 = MAIN, Value = 164.97 P1 = MAIN (V), Value = 6.51 P2 = Yoke, Value = 25.37 P3 = Bottom Coil, Value = 26.95 P4 = Top Coil, Value = 27.92 P5 = Water In, Value = 30.45 P6 = Water Out, Value = 28.59 P7 = Ambient, Value = 23.70 P8 = Supply Pres(psi), Value = 124.19 P9 = Return Pres(psi), Value = 11.57 14:27:20 Magnet Current for MAIN, Imag = 164.9743 A Vtrans = 1.649796 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 14:27:20 MAIN (V) Voltage Measurement, V = 6.509800 V 14:27:24 Temperature Measurement, Probe Number = 0, T = 25.389000 C 14:27:25 Temperature Measurement, Probe Number = 1, T = 24.922000 C 14:27:25 Temperature Measurement, Probe Number = 2, T = 28.132000 C 14:27:26 Temperature Measurement, Probe Number = 3, T = 30.418000 C 14:27:26 Temperature Measurement, Probe Number = 4, T = 28.554000 C 14:27:27 Temperature Measurement, Probe Number = 5, T = 23.752000 C 14:27:28 Sensor Voltage Measurement, Sensor Number = 0, V = 2.655177 V 14:27:28 Sensor Measurement, Sensor Number = 0, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 2.655177 (V, A, ohm, ...), Measured Value = 124.138298 (C, atm, Hz, ...) 14:27:28 Sensor Voltage Measurement, Sensor Number = 1, V = 1.156294 V 14:27:28 Sensor Measurement, Sensor Number = 1, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 1.156294 (V, A, ohm, ...), Measured Value = 11.722043 (C, atm, Hz, ...) 14:27:28 Magnet Test Parameter Measurements P0 = MAIN, Value = 164.97 P1 = MAIN (V), Value = 6.51 P2 = Yoke, Value = 25.39 P3 = Bottom Coil, Value = 24.92 P4 = Top Coil, Value = 28.13 P5 = Water In, Value = 30.42 P6 = Water Out, Value = 28.55 P7 = Ambient, Value = 23.75 P8 = Supply Pres(psi), Value = 124.14 P9 = Return Pres(psi), Value = 11.72 14:28:29 Magnet Current for MAIN, Imag = 164.9743 A Vtrans = 1.649796 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 14:28:29 MAIN (V) Voltage Measurement, V = 6.509683 V 14:28:33 Temperature Measurement, Probe Number = 0, T = 25.400000 C 14:28:34 Temperature Measurement, Probe Number = 1, T = 24.239000 C 14:28:34 Temperature Measurement, Probe Number = 2, T = 28.032000 C 14:28:35 Temperature Measurement, Probe Number = 3, T = 30.435000 C 14:28:35 Temperature Measurement, Probe Number = 4, T = 28.586000 C 14:28:36 Temperature Measurement, Probe Number = 5, T = 23.666000 C 14:28:36 Sensor Voltage Measurement, Sensor Number = 0, V = 2.652600 V 14:28:36 Sensor Measurement, Sensor Number = 0, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 2.652600 (V, A, ohm, ...), Measured Value = 123.945000 (C, atm, Hz, ...) 14:28:37 Sensor Voltage Measurement, Sensor Number = 1, V = 1.156246 V 14:28:37 Sensor Measurement, Sensor Number = 1, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 1.156246 (V, A, ohm, ...), Measured Value = 11.718480 (C, atm, Hz, ...) 14:28:37 Magnet Test Parameter Measurements P0 = MAIN, Value = 164.97 P1 = MAIN (V), Value = 6.51 P2 = Yoke, Value = 25.40 P3 = Bottom Coil, Value = 24.24 P4 = Top Coil, Value = 28.03 P5 = Water In, Value = 30.43 P6 = Water Out, Value = 28.59 P7 = Ambient, Value = 23.67 P8 = Supply Pres(psi), Value = 123.95 P9 = Return Pres(psi), Value = 11.72 14:29:38 Magnet Current for MAIN, Imag = 164.9747 A Vtrans = 1.649800 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 14:29:38 MAIN (V) Voltage Measurement, V = 6.509704 V 14:29:42 Temperature Measurement, Probe Number = 0, T = 25.410000 C 14:29:43 Temperature Measurement, Probe Number = 1, T = 23.883000 C 14:29:43 Temperature Measurement, Probe Number = 2, T = 27.935000 C 14:29:44 Temperature Measurement, Probe Number = 3, T = 30.428000 C 14:29:44 Temperature Measurement, Probe Number = 4, T = 28.594000 C 14:29:45 Temperature Measurement, Probe Number = 5, T = 23.713000 C 14:29:45 Sensor Voltage Measurement, Sensor Number = 0, V = 2.649742 V 14:29:45 Sensor Measurement, Sensor Number = 0, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 2.649742 (V, A, ohm, ...), Measured Value = 123.730635 (C, atm, Hz, ...) 14:29:46 Sensor Voltage Measurement, Sensor Number = 1, V = 1.156373 V 14:29:46 Sensor Measurement, Sensor Number = 1, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 1.156373 (V, A, ohm, ...), Measured Value = 11.727938 (C, atm, Hz, ...) 14:29:46 Magnet Test Parameter Measurements P0 = MAIN, Value = 164.97 P1 = MAIN (V), Value = 6.51 P2 = Yoke, Value = 25.41 P3 = Bottom Coil, Value = 23.88 P4 = Top Coil, Value = 27.93 P5 = Water In, Value = 30.43 P6 = Water Out, Value = 28.59 P7 = Ambient, Value = 23.71 P8 = Supply Pres(psi), Value = 123.73 P9 = Return Pres(psi), Value = 11.73 14:30:47 Magnet Current for MAIN, Imag = 164.9745 A Vtrans = 1.649799 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 14:30:47 MAIN (V) Voltage Measurement, V = 6.509892 V 14:30:51 Temperature Measurement, Probe Number = 0, T = 25.415000 C 14:30:52 Temperature Measurement, Probe Number = 1, T = 23.728000 C 14:30:52 Temperature Measurement, Probe Number = 2, T = 27.876000 C 14:30:53 Temperature Measurement, Probe Number = 3, T = 30.439000 C 14:30:53 Temperature Measurement, Probe Number = 4, T = 28.616000 C 14:30:54 Temperature Measurement, Probe Number = 5, T = 23.734000 C 14:30:54 Sensor Voltage Measurement, Sensor Number = 0, V = 2.654899 V 14:30:54 Sensor Measurement, Sensor Number = 0, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 2.654899 (V, A, ohm, ...), Measured Value = 124.117410 (C, atm, Hz, ...) 14:30:55 Sensor Voltage Measurement, Sensor Number = 1, V = 1.156075 V 14:30:55 Sensor Measurement, Sensor Number = 1, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 1.156075 (V, A, ohm, ...), Measured Value = 11.705647 (C, atm, Hz, ...) 14:30:55 Magnet Test Parameter Measurements P0 = MAIN, Value = 164.97 P1 = MAIN (V), Value = 6.51 P2 = Yoke, Value = 25.41 P3 = Bottom Coil, Value = 23.73 P4 = Top Coil, Value = 27.88 P5 = Water In, Value = 30.44 P6 = Water Out, Value = 28.62 P7 = Ambient, Value = 23.73 P8 = Supply Pres(psi), Value = 124.12 P9 = Return Pres(psi), Value = 11.71 14:31:56 Magnet Current for MAIN, Imag = 164.9745 A Vtrans = 1.649799 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 14:31:56 MAIN (V) Voltage Measurement, V = 6.509946 V 14:32:00 Temperature Measurement, Probe Number = 0, T = 25.445000 C 14:32:01 Temperature Measurement, Probe Number = 1, T = 23.630000 C 14:32:01 Temperature Measurement, Probe Number = 2, T = 27.884000 C 14:32:02 Temperature Measurement, Probe Number = 3, T = 30.452000 C 14:32:02 Temperature Measurement, Probe Number = 4, T = 28.617000 C 14:32:03 Temperature Measurement, Probe Number = 5, T = 23.774000 C 14:32:03 Sensor Voltage Measurement, Sensor Number = 0, V = 2.663070 V 14:32:03 Sensor Measurement, Sensor Number = 0, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 2.663070 (V, A, ohm, ...), Measured Value = 124.730258 (C, atm, Hz, ...) 14:32:04 Sensor Voltage Measurement, Sensor Number = 1, V = 1.157641 V 14:32:04 Sensor Measurement, Sensor Number = 1, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 1.157641 (V, A, ohm, ...), Measured Value = 11.823090 (C, atm, Hz, ...) 14:32:04 Magnet Test Parameter Measurements P0 = MAIN, Value = 164.97 P1 = MAIN (V), Value = 6.51 P2 = Yoke, Value = 25.45 P3 = Bottom Coil, Value = 23.63 P4 = Top Coil, Value = 27.88 P5 = Water In, Value = 30.45 P6 = Water Out, Value = 28.62 P7 = Ambient, Value = 23.77 P8 = Supply Pres(psi), Value = 124.73 P9 = Return Pres(psi), Value = 11.82 14:33:05 Magnet Current for MAIN, Imag = 164.9743 A Vtrans = 1.649796 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 14:33:05 MAIN (V) Voltage Measurement, V = 6.510206 V 14:33:09 Temperature Measurement, Probe Number = 0, T = 25.448000 C 14:33:10 Temperature Measurement, Probe Number = 1, T = 23.685000 C 14:33:10 Temperature Measurement, Probe Number = 2, T = 27.913000 C 14:33:11 Temperature Measurement, Probe Number = 3, T = 30.458000 C 14:33:11 Temperature Measurement, Probe Number = 4, T = 28.622000 C 14:33:12 Temperature Measurement, Probe Number = 5, T = 23.810000 C 14:33:12 Sensor Voltage Measurement, Sensor Number = 0, V = 2.667301 V 14:33:12 Sensor Measurement, Sensor Number = 0, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 2.667301 (V, A, ohm, ...), Measured Value = 125.047545 (C, atm, Hz, ...) 14:33:13 Sensor Voltage Measurement, Sensor Number = 1, V = 1.154579 V 14:33:13 Sensor Measurement, Sensor Number = 1, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 1.154579 (V, A, ohm, ...), Measured Value = 11.593455 (C, atm, Hz, ...) 14:33:13 Magnet Test Parameter Measurements P0 = MAIN, Value = 164.97 P1 = MAIN (V), Value = 6.51 P2 = Yoke, Value = 25.45 P3 = Bottom Coil, Value = 23.68 P4 = Top Coil, Value = 27.91 P5 = Water In, Value = 30.46 P6 = Water Out, Value = 28.62 P7 = Ambient, Value = 23.81 P8 = Supply Pres(psi), Value = 125.05 P9 = Return Pres(psi), Value = 11.59 14:34:14 Magnet Current for MAIN, Imag = 164.9744 A Vtrans = 1.649798 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 14:34:14 MAIN (V) Voltage Measurement, V = 6.510327 V 14:34:18 Temperature Measurement, Probe Number = 0, T = 25.459000 C 14:34:19 Temperature Measurement, Probe Number = 1, T = 23.668000 C 14:34:19 Temperature Measurement, Probe Number = 2, T = 27.954000 C 14:34:20 Temperature Measurement, Probe Number = 3, T = 30.464000 C 14:34:20 Temperature Measurement, Probe Number = 4, T = 28.640000 C 14:34:21 Temperature Measurement, Probe Number = 5, T = 23.870000 C 14:34:21 Sensor Voltage Measurement, Sensor Number = 0, V = 2.659729 V 14:34:21 Sensor Measurement, Sensor Number = 0, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 2.659729 (V, A, ohm, ...), Measured Value = 124.479653 (C, atm, Hz, ...) 14:34:22 Sensor Voltage Measurement, Sensor Number = 1, V = 1.154363 V 14:34:22 Sensor Measurement, Sensor Number = 1, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 1.154363 (V, A, ohm, ...), Measured Value = 11.577255 (C, atm, Hz, ...) 14:34:22 Magnet Test Parameter Measurements P0 = MAIN, Value = 164.97 P1 = MAIN (V), Value = 6.51 P2 = Yoke, Value = 25.46 P3 = Bottom Coil, Value = 23.67 P4 = Top Coil, Value = 27.95 P5 = Water In, Value = 30.46 P6 = Water Out, Value = 28.64 P7 = Ambient, Value = 23.87 P8 = Supply Pres(psi), Value = 124.48 P9 = Return Pres(psi), Value = 11.58 14:35:23 Magnet Current for MAIN, Imag = 164.9748 A Vtrans = 1.649801 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 14:35:23 MAIN (V) Voltage Measurement, V = 6.510306 V 14:35:27 Temperature Measurement, Probe Number = 0, T = 25.473000 C 14:35:28 Temperature Measurement, Probe Number = 1, T = 23.672000 C 14:35:28 Temperature Measurement, Probe Number = 2, T = 28.199000 C 14:35:29 Temperature Measurement, Probe Number = 3, T = 30.484000 C 14:35:29 Temperature Measurement, Probe Number = 4, T = 28.618000 C 14:35:30 Temperature Measurement, Probe Number = 5, T = 23.889000 C 14:35:31 Sensor Voltage Measurement, Sensor Number = 0, V = 2.644384 V 14:35:31 Sensor Measurement, Sensor Number = 0, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 2.644384 (V, A, ohm, ...), Measured Value = 123.328785 (C, atm, Hz, ...) 14:35:31 Sensor Voltage Measurement, Sensor Number = 1, V = 1.156606 V 14:35:31 Sensor Measurement, Sensor Number = 1, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 1.156606 (V, A, ohm, ...), Measured Value = 11.745450 (C, atm, Hz, ...) 14:35:31 Magnet Test Parameter Measurements P0 = MAIN, Value = 164.97 P1 = MAIN (V), Value = 6.51 P2 = Yoke, Value = 25.47 P3 = Bottom Coil, Value = 23.67 P4 = Top Coil, Value = 28.20 P5 = Water In, Value = 30.48 P6 = Water Out, Value = 28.62 P7 = Ambient, Value = 23.89 P8 = Supply Pres(psi), Value = 123.33 P9 = Return Pres(psi), Value = 11.75 14:36:32 Magnet Current for MAIN, Imag = 164.9742 A Vtrans = 1.649795 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 14:36:32 MAIN (V) Voltage Measurement, V = 6.510324 V 14:36:36 Temperature Measurement, Probe Number = 0, T = 25.480000 C 14:36:37 Temperature Measurement, Probe Number = 1, T = 23.679000 C 14:36:37 Temperature Measurement, Probe Number = 2, T = 28.050000 C 14:36:38 Temperature Measurement, Probe Number = 3, T = 30.475000 C 14:36:38 Temperature Measurement, Probe Number = 4, T = 28.621000 C 14:36:39 Temperature Measurement, Probe Number = 5, T = 23.837000 C 14:36:40 Sensor Voltage Measurement, Sensor Number = 0, V = 2.653759 V 14:36:40 Sensor Measurement, Sensor Number = 0, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 2.653759 (V, A, ohm, ...), Measured Value = 124.031910 (C, atm, Hz, ...) 14:36:40 Sensor Voltage Measurement, Sensor Number = 1, V = 1.154639 V 14:36:40 Sensor Measurement, Sensor Number = 1, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 1.154639 (V, A, ohm, ...), Measured Value = 11.597955 (C, atm, Hz, ...) 14:36:40 Magnet Test Parameter Measurements P0 = MAIN, Value = 164.97 P1 = MAIN (V), Value = 6.51 P2 = Yoke, Value = 25.48 P3 = Bottom Coil, Value = 23.68 P4 = Top Coil, Value = 28.05 P5 = Water In, Value = 30.48 P6 = Water Out, Value = 28.62 P7 = Ambient, Value = 23.84 P8 = Supply Pres(psi), Value = 124.03 P9 = Return Pres(psi), Value = 11.60 14:37:41 Magnet Current for MAIN, Imag = 164.9745 A Vtrans = 1.649799 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 14:37:41 MAIN (V) Voltage Measurement, V = 6.510484 V 14:37:45 Temperature Measurement, Probe Number = 0, T = 25.485000 C 14:37:46 Temperature Measurement, Probe Number = 1, T = 23.731000 C 14:37:46 Temperature Measurement, Probe Number = 2, T = 28.034000 C 14:37:47 Temperature Measurement, Probe Number = 3, T = 30.471000 C 14:37:47 Temperature Measurement, Probe Number = 4, T = 28.600000 C 14:37:48 Temperature Measurement, Probe Number = 5, T = 23.866000 C 14:37:49 Sensor Voltage Measurement, Sensor Number = 0, V = 2.662769 V 14:37:49 Sensor Measurement, Sensor Number = 0, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 2.662769 (V, A, ohm, ...), Measured Value = 124.707690 (C, atm, Hz, ...) 14:37:49 Sensor Voltage Measurement, Sensor Number = 1, V = 1.158099 V 14:37:49 Sensor Measurement, Sensor Number = 1, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 1.158099 (V, A, ohm, ...), Measured Value = 11.857455 (C, atm, Hz, ...) 14:37:49 Magnet Test Parameter Measurements P0 = MAIN, Value = 164.97 P1 = MAIN (V), Value = 6.51 P2 = Yoke, Value = 25.48 P3 = Bottom Coil, Value = 23.73 P4 = Top Coil, Value = 28.03 P5 = Water In, Value = 30.47 P6 = Water Out, Value = 28.60 P7 = Ambient, Value = 23.87 P8 = Supply Pres(psi), Value = 124.71 P9 = Return Pres(psi), Value = 11.86 14:38:50 Magnet Current for MAIN, Imag = 164.9747 A Vtrans = 1.649800 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 14:38:50 MAIN (V) Voltage Measurement, V = 6.510313 V 14:38:54 Temperature Measurement, Probe Number = 0, T = 25.488000 C 14:38:55 Temperature Measurement, Probe Number = 1, T = 23.706000 C 14:38:55 Temperature Measurement, Probe Number = 2, T = 28.011000 C 14:38:56 Temperature Measurement, Probe Number = 3, T = 30.458000 C 14:38:56 Temperature Measurement, Probe Number = 4, T = 28.557000 C 14:38:57 Temperature Measurement, Probe Number = 5, T = 23.849000 C 14:38:58 Sensor Voltage Measurement, Sensor Number = 0, V = 2.658463 V 14:38:58 Sensor Measurement, Sensor Number = 0, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 2.658463 (V, A, ohm, ...), Measured Value = 124.384688 (C, atm, Hz, ...) 14:38:58 Sensor Voltage Measurement, Sensor Number = 1, V = 1.157126 V 14:38:58 Sensor Measurement, Sensor Number = 1, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 1.157126 (V, A, ohm, ...), Measured Value = 11.784413 (C, atm, Hz, ...) 14:38:58 Magnet Test Parameter Measurements P0 = MAIN, Value = 164.97 P1 = MAIN (V), Value = 6.51 P2 = Yoke, Value = 25.49 P3 = Bottom Coil, Value = 23.71 P4 = Top Coil, Value = 28.01 P5 = Water In, Value = 30.46 P6 = Water Out, Value = 28.56 P7 = Ambient, Value = 23.85 P8 = Supply Pres(psi), Value = 124.38 P9 = Return Pres(psi), Value = 11.78 14:39:59 Magnet Current for MAIN, Imag = 164.9747 A Vtrans = 1.649800 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 14:39:59 MAIN (V) Voltage Measurement, V = 6.510383 V 14:40:03 Temperature Measurement, Probe Number = 0, T = 25.495000 C 14:40:04 Temperature Measurement, Probe Number = 1, T = 23.698000 C 14:40:04 Temperature Measurement, Probe Number = 2, T = 28.047000 C 14:40:05 Temperature Measurement, Probe Number = 3, T = 30.445000 C 14:40:05 Temperature Measurement, Probe Number = 4, T = 28.587000 C 14:40:06 Temperature Measurement, Probe Number = 5, T = 23.857000 C 14:40:07 Sensor Voltage Measurement, Sensor Number = 0, V = 2.654025 V 14:40:07 Sensor Measurement, Sensor Number = 0, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 2.654025 (V, A, ohm, ...), Measured Value = 124.051860 (C, atm, Hz, ...) 14:40:07 Sensor Voltage Measurement, Sensor Number = 1, V = 1.156146 V 14:40:07 Sensor Measurement, Sensor Number = 1, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 1.156146 (V, A, ohm, ...), Measured Value = 11.710987 (C, atm, Hz, ...) 14:40:07 Magnet Test Parameter Measurements P0 = MAIN, Value = 164.97 P1 = MAIN (V), Value = 6.51 P2 = Yoke, Value = 25.50 P3 = Bottom Coil, Value = 23.70 P4 = Top Coil, Value = 28.05 P5 = Water In, Value = 30.45 P6 = Water Out, Value = 28.59 P7 = Ambient, Value = 23.86 P8 = Supply Pres(psi), Value = 124.05 P9 = Return Pres(psi), Value = 11.71 14:41:08 Magnet Current for MAIN, Imag = 164.9747 A Vtrans = 1.649800 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 14:41:08 MAIN (V) Voltage Measurement, V = 6.510303 V 14:41:12 Temperature Measurement, Probe Number = 0, T = 25.503000 C 14:41:13 Temperature Measurement, Probe Number = 1, T = 23.717000 C 14:41:13 Temperature Measurement, Probe Number = 2, T = 28.019000 C 14:41:14 Temperature Measurement, Probe Number = 3, T = 30.452000 C 14:41:14 Temperature Measurement, Probe Number = 4, T = 28.569000 C 14:41:15 Temperature Measurement, Probe Number = 5, T = 23.869000 C 14:41:16 Sensor Voltage Measurement, Sensor Number = 0, V = 2.639751 V 14:41:16 Sensor Measurement, Sensor Number = 0, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 2.639751 (V, A, ohm, ...), Measured Value = 122.981347 (C, atm, Hz, ...) 14:41:16 Sensor Voltage Measurement, Sensor Number = 1, V = 1.155341 V 14:41:16 Sensor Measurement, Sensor Number = 1, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 1.155341 (V, A, ohm, ...), Measured Value = 11.650583 (C, atm, Hz, ...) 14:41:16 Magnet Test Parameter Measurements P0 = MAIN, Value = 164.97 P1 = MAIN (V), Value = 6.51 P2 = Yoke, Value = 25.50 P3 = Bottom Coil, Value = 23.72 P4 = Top Coil, Value = 28.02 P5 = Water In, Value = 30.45 P6 = Water Out, Value = 28.57 P7 = Ambient, Value = 23.87 P8 = Supply Pres(psi), Value = 122.98 P9 = Return Pres(psi), Value = 11.65 14:42:17 Magnet Current for MAIN, Imag = 164.9747 A Vtrans = 1.649800 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 14:42:18 MAIN (V) Voltage Measurement, V = 6.510573 V 14:42:21 Temperature Measurement, Probe Number = 0, T = 25.505000 C 14:42:22 Temperature Measurement, Probe Number = 1, T = 23.747000 C 14:42:22 Temperature Measurement, Probe Number = 2, T = 28.147000 C 14:42:23 Temperature Measurement, Probe Number = 3, T = 30.466000 C 14:42:23 Temperature Measurement, Probe Number = 4, T = 28.601000 C 14:42:24 Temperature Measurement, Probe Number = 5, T = 23.877000 C 14:42:25 Sensor Voltage Measurement, Sensor Number = 0, V = 2.662822 V 14:42:25 Sensor Measurement, Sensor Number = 0, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 2.662822 (V, A, ohm, ...), Measured Value = 124.711620 (C, atm, Hz, ...) 14:42:25 Sensor Voltage Measurement, Sensor Number = 1, V = 1.154195 V 14:42:25 Sensor Measurement, Sensor Number = 1, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 1.154195 (V, A, ohm, ...), Measured Value = 11.564610 (C, atm, Hz, ...) 14:42:25 Magnet Test Parameter Measurements P0 = MAIN, Value = 164.97 P1 = MAIN (V), Value = 6.51 P2 = Yoke, Value = 25.50 P3 = Bottom Coil, Value = 23.75 P4 = Top Coil, Value = 28.15 P5 = Water In, Value = 30.47 P6 = Water Out, Value = 28.60 P7 = Ambient, Value = 23.88 P8 = Supply Pres(psi), Value = 124.71 P9 = Return Pres(psi), Value = 11.56 14:43:26 Magnet Current for MAIN, Imag = 164.9749 A Vtrans = 1.649803 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 14:43:27 MAIN (V) Voltage Measurement, V = 6.510712 V 14:43:30 Temperature Measurement, Probe Number = 0, T = 25.525000 C 14:43:31 Temperature Measurement, Probe Number = 1, T = 23.752000 C 14:43:31 Temperature Measurement, Probe Number = 2, T = 28.136000 C 14:43:32 Temperature Measurement, Probe Number = 3, T = 30.465000 C 14:43:32 Temperature Measurement, Probe Number = 4, T = 28.600000 C 14:43:33 Temperature Measurement, Probe Number = 5, T = 23.876000 C 14:43:34 Sensor Voltage Measurement, Sensor Number = 0, V = 2.653412 V 14:43:34 Sensor Measurement, Sensor Number = 0, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 2.653412 (V, A, ohm, ...), Measured Value = 124.005877 (C, atm, Hz, ...) 14:43:34 Sensor Voltage Measurement, Sensor Number = 1, V = 1.157672 V 14:43:34 Sensor Measurement, Sensor Number = 1, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 1.157672 (V, A, ohm, ...), Measured Value = 11.825430 (C, atm, Hz, ...) 14:43:34 Magnet Test Parameter Measurements P0 = MAIN, Value = 164.97 P1 = MAIN (V), Value = 6.51 P2 = Yoke, Value = 25.52 P3 = Bottom Coil, Value = 23.75 P4 = Top Coil, Value = 28.14 P5 = Water In, Value = 30.46 P6 = Water Out, Value = 28.60 P7 = Ambient, Value = 23.88 P8 = Supply Pres(psi), Value = 124.01 P9 = Return Pres(psi), Value = 11.83 14:44:35 Magnet Current for MAIN, Imag = 164.9748 A Vtrans = 1.649801 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 14:44:36 MAIN (V) Voltage Measurement, V = 6.510758 V 14:44:39 Temperature Measurement, Probe Number = 0, T = 25.527000 C 14:44:40 Temperature Measurement, Probe Number = 1, T = 23.769000 C 14:44:40 Temperature Measurement, Probe Number = 2, T = 28.045000 C 14:44:41 Temperature Measurement, Probe Number = 3, T = 30.468000 C 14:44:41 Temperature Measurement, Probe Number = 4, T = 28.625000 C 14:44:42 Temperature Measurement, Probe Number = 5, T = 23.877000 C 14:44:43 Sensor Voltage Measurement, Sensor Number = 0, V = 2.657687 V 14:44:43 Sensor Measurement, Sensor Number = 0, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 2.657687 (V, A, ohm, ...), Measured Value = 124.326532 (C, atm, Hz, ...) 14:44:43 Sensor Voltage Measurement, Sensor Number = 1, V = 1.154510 V 14:44:43 Sensor Measurement, Sensor Number = 1, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 1.154510 (V, A, ohm, ...), Measured Value = 11.588213 (C, atm, Hz, ...) 14:44:43 Magnet Test Parameter Measurements P0 = MAIN, Value = 164.97 P1 = MAIN (V), Value = 6.51 P2 = Yoke, Value = 25.53 P3 = Bottom Coil, Value = 23.77 P4 = Top Coil, Value = 28.05 P5 = Water In, Value = 30.47 P6 = Water Out, Value = 28.62 P7 = Ambient, Value = 23.88 P8 = Supply Pres(psi), Value = 124.33 P9 = Return Pres(psi), Value = 11.59 14:45:44 Magnet Current for MAIN, Imag = 164.9745 A Vtrans = 1.649799 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 14:45:45 MAIN (V) Voltage Measurement, V = 6.510851 V 14:45:48 Temperature Measurement, Probe Number = 0, T = 25.535000 C 14:45:49 Temperature Measurement, Probe Number = 1, T = 23.759000 C 14:45:49 Temperature Measurement, Probe Number = 2, T = 28.010000 C 14:45:50 Temperature Measurement, Probe Number = 3, T = 30.487000 C 14:45:50 Temperature Measurement, Probe Number = 4, T = 28.629000 C 14:45:51 Temperature Measurement, Probe Number = 5, T = 23.906000 C 14:45:52 Sensor Voltage Measurement, Sensor Number = 0, V = 2.640210 V 14:45:52 Sensor Measurement, Sensor Number = 0, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 2.640210 (V, A, ohm, ...), Measured Value = 123.015713 (C, atm, Hz, ...) 14:45:52 Sensor Voltage Measurement, Sensor Number = 1, V = 1.154995 V 14:45:52 Sensor Measurement, Sensor Number = 1, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 1.154995 (V, A, ohm, ...), Measured Value = 11.624640 (C, atm, Hz, ...) 14:45:52 Magnet Test Parameter Measurements P0 = MAIN, Value = 164.97 P1 = MAIN (V), Value = 6.51 P2 = Yoke, Value = 25.54 P3 = Bottom Coil, Value = 23.76 P4 = Top Coil, Value = 28.01 P5 = Water In, Value = 30.49 P6 = Water Out, Value = 28.63 P7 = Ambient, Value = 23.91 P8 = Supply Pres(psi), Value = 123.02 P9 = Return Pres(psi), Value = 11.62 14:46:53 Magnet Current for MAIN, Imag = 164.9752 A Vtrans = 1.649805 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 14:46:54 MAIN (V) Voltage Measurement, V = 6.511044 V 14:46:57 Temperature Measurement, Probe Number = 0, T = 25.548000 C 14:46:58 Temperature Measurement, Probe Number = 1, T = 23.789000 C 14:46:58 Temperature Measurement, Probe Number = 2, T = 27.995000 C 14:46:59 Temperature Measurement, Probe Number = 3, T = 30.496000 C 14:47:00 Temperature Measurement, Probe Number = 4, T = 28.658000 C 14:47:00 Temperature Measurement, Probe Number = 5, T = 23.912000 C 14:47:01 Sensor Voltage Measurement, Sensor Number = 0, V = 2.667564 V 14:47:01 Sensor Measurement, Sensor Number = 0, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 2.667564 (V, A, ohm, ...), Measured Value = 125.067300 (C, atm, Hz, ...) 14:47:01 Sensor Voltage Measurement, Sensor Number = 1, V = 1.158183 V 14:47:01 Sensor Measurement, Sensor Number = 1, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 1.158183 (V, A, ohm, ...), Measured Value = 11.863733 (C, atm, Hz, ...) 14:47:01 Magnet Test Parameter Measurements P0 = MAIN, Value = 164.98 P1 = MAIN (V), Value = 6.51 P2 = Yoke, Value = 25.55 P3 = Bottom Coil, Value = 23.79 P4 = Top Coil, Value = 28.00 P5 = Water In, Value = 30.50 P6 = Water Out, Value = 28.66 P7 = Ambient, Value = 23.91 P8 = Supply Pres(psi), Value = 125.07 P9 = Return Pres(psi), Value = 11.86 14:48:02 Magnet Current for MAIN, Imag = 164.9749 A Vtrans = 1.649803 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 14:48:03 MAIN (V) Voltage Measurement, V = 6.511062 V 14:48:06 Temperature Measurement, Probe Number = 0, T = 25.554000 C 14:48:07 Temperature Measurement, Probe Number = 1, T = 23.764000 C 14:48:07 Temperature Measurement, Probe Number = 2, T = 28.130000 C 14:48:08 Temperature Measurement, Probe Number = 3, T = 30.497000 C 14:48:09 Temperature Measurement, Probe Number = 4, T = 28.655000 C 14:48:09 Temperature Measurement, Probe Number = 5, T = 23.921000 C 14:48:10 Sensor Voltage Measurement, Sensor Number = 0, V = 2.641950 V 14:48:10 Sensor Measurement, Sensor Number = 0, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 2.641950 (V, A, ohm, ...), Measured Value = 123.146265 (C, atm, Hz, ...) 14:48:10 Sensor Voltage Measurement, Sensor Number = 1, V = 1.154659 V 14:48:10 Sensor Measurement, Sensor Number = 1, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 1.154659 (V, A, ohm, ...), Measured Value = 11.599455 (C, atm, Hz, ...) 14:48:10 Magnet Test Parameter Measurements P0 = MAIN, Value = 164.97 P1 = MAIN (V), Value = 6.51 P2 = Yoke, Value = 25.55 P3 = Bottom Coil, Value = 23.76 P4 = Top Coil, Value = 28.13 P5 = Water In, Value = 30.50 P6 = Water Out, Value = 28.66 P7 = Ambient, Value = 23.92 P8 = Supply Pres(psi), Value = 123.15 P9 = Return Pres(psi), Value = 11.60 14:49:11 Magnet Current for MAIN, Imag = 164.9749 A Vtrans = 1.649803 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 14:49:12 MAIN (V) Voltage Measurement, V = 6.511060 V 14:49:15 Temperature Measurement, Probe Number = 0, T = 25.572000 C 14:49:16 Temperature Measurement, Probe Number = 1, T = 23.717000 C 14:49:16 Temperature Measurement, Probe Number = 2, T = 28.015000 C 14:49:17 Temperature Measurement, Probe Number = 3, T = 30.502000 C 14:49:18 Temperature Measurement, Probe Number = 4, T = 28.661000 C 14:49:18 Temperature Measurement, Probe Number = 5, T = 23.940000 C 14:49:19 Sensor Voltage Measurement, Sensor Number = 0, V = 2.650110 V 14:49:19 Sensor Measurement, Sensor Number = 0, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 2.650110 (V, A, ohm, ...), Measured Value = 123.758265 (C, atm, Hz, ...) 14:49:19 Sensor Voltage Measurement, Sensor Number = 1, V = 1.155496 V 14:49:19 Sensor Measurement, Sensor Number = 1, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 1.155496 (V, A, ohm, ...), Measured Value = 11.662200 (C, atm, Hz, ...) 14:49:19 Magnet Test Parameter Measurements P0 = MAIN, Value = 164.97 P1 = MAIN (V), Value = 6.51 P2 = Yoke, Value = 25.57 P3 = Bottom Coil, Value = 23.72 P4 = Top Coil, Value = 28.02 P5 = Water In, Value = 30.50 P6 = Water Out, Value = 28.66 P7 = Ambient, Value = 23.94 P8 = Supply Pres(psi), Value = 123.76 P9 = Return Pres(psi), Value = 11.66 14:50:20 Magnet Current for MAIN, Imag = 164.9749 A Vtrans = 1.649803 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 14:50:21 MAIN (V) Voltage Measurement, V = 6.511216 V 14:50:24 Temperature Measurement, Probe Number = 0, T = 25.568000 C 14:50:25 Temperature Measurement, Probe Number = 1, T = 23.773000 C 14:50:25 Temperature Measurement, Probe Number = 2, T = 28.087000 C 14:50:26 Temperature Measurement, Probe Number = 3, T = 30.507000 C 14:50:27 Temperature Measurement, Probe Number = 4, T = 28.652000 C 14:50:27 Temperature Measurement, Probe Number = 5, T = 23.927000 C 14:50:28 Sensor Voltage Measurement, Sensor Number = 0, V = 2.644818 V 14:50:28 Sensor Measurement, Sensor Number = 0, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 2.644818 (V, A, ohm, ...), Measured Value = 123.361372 (C, atm, Hz, ...) 14:50:28 Sensor Voltage Measurement, Sensor Number = 1, V = 1.155656 V 14:50:28 Sensor Measurement, Sensor Number = 1, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 1.155656 (V, A, ohm, ...), Measured Value = 11.674185 (C, atm, Hz, ...) 14:50:28 Magnet Test Parameter Measurements P0 = MAIN, Value = 164.97 P1 = MAIN (V), Value = 6.51 P2 = Yoke, Value = 25.57 P3 = Bottom Coil, Value = 23.77 P4 = Top Coil, Value = 28.09 P5 = Water In, Value = 30.51 P6 = Water Out, Value = 28.65 P7 = Ambient, Value = 23.93 P8 = Supply Pres(psi), Value = 123.36 P9 = Return Pres(psi), Value = 11.67 14:51:29 Magnet Current for MAIN, Imag = 164.9752 A Vtrans = 1.649805 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 14:51:30 MAIN (V) Voltage Measurement, V = 6.511327 V 14:51:33 Temperature Measurement, Probe Number = 0, T = 25.566000 C 14:51:34 Temperature Measurement, Probe Number = 1, T = 23.719000 C 14:51:34 Temperature Measurement, Probe Number = 2, T = 28.012000 C 14:51:35 Temperature Measurement, Probe Number = 3, T = 30.515000 C 14:51:36 Temperature Measurement, Probe Number = 4, T = 28.674000 C 14:51:36 Temperature Measurement, Probe Number = 5, T = 23.961000 C 14:51:37 Sensor Voltage Measurement, Sensor Number = 0, V = 2.655709 V 14:51:37 Sensor Measurement, Sensor Number = 0, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 2.655709 (V, A, ohm, ...), Measured Value = 124.178190 (C, atm, Hz, ...) 14:51:37 Sensor Voltage Measurement, Sensor Number = 1, V = 1.153863 V 14:51:37 Sensor Measurement, Sensor Number = 1, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 1.153863 (V, A, ohm, ...), Measured Value = 11.539702 (C, atm, Hz, ...) 14:51:37 Magnet Test Parameter Measurements P0 = MAIN, Value = 164.98 P1 = MAIN (V), Value = 6.51 P2 = Yoke, Value = 25.57 P3 = Bottom Coil, Value = 23.72 P4 = Top Coil, Value = 28.01 P5 = Water In, Value = 30.52 P6 = Water Out, Value = 28.67 P7 = Ambient, Value = 23.96 P8 = Supply Pres(psi), Value = 124.18 P9 = Return Pres(psi), Value = 11.54 14:52:38 Magnet Current for MAIN, Imag = 164.9750 A Vtrans = 1.649804 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 14:52:39 MAIN (V) Voltage Measurement, V = 6.511394 V 14:52:42 Temperature Measurement, Probe Number = 0, T = 25.579000 C 14:52:43 Temperature Measurement, Probe Number = 1, T = 23.767000 C 14:52:43 Temperature Measurement, Probe Number = 2, T = 27.995000 C 14:52:44 Temperature Measurement, Probe Number = 3, T = 30.520000 C 14:52:45 Temperature Measurement, Probe Number = 4, T = 28.677000 C 14:52:45 Temperature Measurement, Probe Number = 5, T = 23.969000 C 14:52:46 Sensor Voltage Measurement, Sensor Number = 0, V = 2.649413 V 14:52:46 Sensor Measurement, Sensor Number = 0, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 2.649413 (V, A, ohm, ...), Measured Value = 123.706005 (C, atm, Hz, ...) 14:52:46 Sensor Voltage Measurement, Sensor Number = 1, V = 1.155018 V 14:52:46 Sensor Measurement, Sensor Number = 1, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 1.155018 (V, A, ohm, ...), Measured Value = 11.626327 (C, atm, Hz, ...) 14:52:46 Magnet Test Parameter Measurements P0 = MAIN, Value = 164.98 P1 = MAIN (V), Value = 6.51 P2 = Yoke, Value = 25.58 P3 = Bottom Coil, Value = 23.77 P4 = Top Coil, Value = 28.00 P5 = Water In, Value = 30.52 P6 = Water Out, Value = 28.68 P7 = Ambient, Value = 23.97 P8 = Supply Pres(psi), Value = 123.71 P9 = Return Pres(psi), Value = 11.63 14:53:47 Magnet Current for MAIN, Imag = 164.9755 A Vtrans = 1.649809 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 14:53:48 MAIN (V) Voltage Measurement, V = 6.511297 V 14:53:51 Temperature Measurement, Probe Number = 0, T = 25.594000 C 14:53:52 Temperature Measurement, Probe Number = 1, T = 23.754000 C 14:53:52 Temperature Measurement, Probe Number = 2, T = 27.988000 C 14:53:53 Temperature Measurement, Probe Number = 3, T = 30.524000 C 14:53:54 Temperature Measurement, Probe Number = 4, T = 28.692000 C 14:53:54 Temperature Measurement, Probe Number = 5, T = 24.004000 C 14:53:55 Sensor Voltage Measurement, Sensor Number = 0, V = 2.657064 V 14:53:55 Sensor Measurement, Sensor Number = 0, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 2.657064 (V, A, ohm, ...), Measured Value = 124.279800 (C, atm, Hz, ...) 14:53:55 Sensor Voltage Measurement, Sensor Number = 1, V = 1.157466 V 14:53:55 Sensor Measurement, Sensor Number = 1, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 1.157466 (V, A, ohm, ...), Measured Value = 11.809980 (C, atm, Hz, ...) 14:53:55 Magnet Test Parameter Measurements P0 = MAIN, Value = 164.98 P1 = MAIN (V), Value = 6.51 P2 = Yoke, Value = 25.59 P3 = Bottom Coil, Value = 23.75 P4 = Top Coil, Value = 27.99 P5 = Water In, Value = 30.52 P6 = Water Out, Value = 28.69 P7 = Ambient, Value = 24.00 P8 = Supply Pres(psi), Value = 124.28 P9 = Return Pres(psi), Value = 11.81 14:54:56 Magnet Current for MAIN, Imag = 164.9755 A Vtrans = 1.649809 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 14:54:57 MAIN (V) Voltage Measurement, V = 6.511346 V 14:55:00 Temperature Measurement, Probe Number = 0, T = 25.601000 C 14:55:01 Temperature Measurement, Probe Number = 1, T = 23.770000 C 14:55:01 Temperature Measurement, Probe Number = 2, T = 28.025000 C 14:55:02 Temperature Measurement, Probe Number = 3, T = 30.525000 C 14:55:03 Temperature Measurement, Probe Number = 4, T = 28.701000 C 14:55:03 Temperature Measurement, Probe Number = 5, T = 24.006000 C 14:55:04 Sensor Voltage Measurement, Sensor Number = 0, V = 2.663904 V 14:55:04 Sensor Measurement, Sensor Number = 0, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 2.663904 (V, A, ohm, ...), Measured Value = 124.792815 (C, atm, Hz, ...) 14:55:04 Sensor Voltage Measurement, Sensor Number = 1, V = 1.160402 V 14:55:04 Sensor Measurement, Sensor Number = 1, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 1.160402 (V, A, ohm, ...), Measured Value = 12.030150 (C, atm, Hz, ...) 14:55:04 Magnet Test Parameter Measurements P0 = MAIN, Value = 164.98 P1 = MAIN (V), Value = 6.51 P2 = Yoke, Value = 25.60 P3 = Bottom Coil, Value = 23.77 P4 = Top Coil, Value = 28.02 P5 = Water In, Value = 30.52 P6 = Water Out, Value = 28.70 P7 = Ambient, Value = 24.01 P8 = Supply Pres(psi), Value = 124.79 P9 = Return Pres(psi), Value = 12.03 14:56:05 Magnet Current for MAIN, Imag = 164.9750 A Vtrans = 1.649804 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 14:56:06 MAIN (V) Voltage Measurement, V = 6.511380 V 14:56:09 Temperature Measurement, Probe Number = 0, T = 25.594000 C 14:56:10 Temperature Measurement, Probe Number = 1, T = 23.773000 C 14:56:10 Temperature Measurement, Probe Number = 2, T = 28.016000 C 14:56:11 Temperature Measurement, Probe Number = 3, T = 30.543000 C 14:56:12 Temperature Measurement, Probe Number = 4, T = 28.661000 C 14:56:12 Temperature Measurement, Probe Number = 5, T = 24.000000 C 14:56:13 Sensor Voltage Measurement, Sensor Number = 0, V = 2.642472 V 14:56:13 Sensor Measurement, Sensor Number = 0, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 2.642472 (V, A, ohm, ...), Measured Value = 123.185407 (C, atm, Hz, ...) 14:56:13 Sensor Voltage Measurement, Sensor Number = 1, V = 1.154953 V 14:56:13 Sensor Measurement, Sensor Number = 1, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 1.154953 (V, A, ohm, ...), Measured Value = 11.621460 (C, atm, Hz, ...) 14:56:13 Magnet Test Parameter Measurements P0 = MAIN, Value = 164.98 P1 = MAIN (V), Value = 6.51 P2 = Yoke, Value = 25.59 P3 = Bottom Coil, Value = 23.77 P4 = Top Coil, Value = 28.02 P5 = Water In, Value = 30.54 P6 = Water Out, Value = 28.66 P7 = Ambient, Value = 24.00 P8 = Supply Pres(psi), Value = 123.19 P9 = Return Pres(psi), Value = 11.62 14:57:14 Magnet Current for MAIN, Imag = 164.9754 A Vtrans = 1.649808 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 14:57:15 MAIN (V) Voltage Measurement, V = 6.511265 V 14:57:18 Temperature Measurement, Probe Number = 0, T = 25.590000 C 14:57:19 Temperature Measurement, Probe Number = 1, T = 23.779000 C 14:57:19 Temperature Measurement, Probe Number = 2, T = 28.128000 C 14:57:20 Temperature Measurement, Probe Number = 3, T = 30.529000 C 14:57:21 Temperature Measurement, Probe Number = 4, T = 28.685000 C 14:57:21 Temperature Measurement, Probe Number = 5, T = 24.033000 C 14:57:22 Sensor Voltage Measurement, Sensor Number = 0, V = 2.644680 V 14:57:22 Sensor Measurement, Sensor Number = 0, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 2.644680 (V, A, ohm, ...), Measured Value = 123.350978 (C, atm, Hz, ...) 14:57:22 Sensor Voltage Measurement, Sensor Number = 1, V = 1.155990 V 14:57:22 Sensor Measurement, Sensor Number = 1, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 1.155990 (V, A, ohm, ...), Measured Value = 11.699280 (C, atm, Hz, ...) 14:57:22 Magnet Test Parameter Measurements P0 = MAIN, Value = 164.98 P1 = MAIN (V), Value = 6.51 P2 = Yoke, Value = 25.59 P3 = Bottom Coil, Value = 23.78 P4 = Top Coil, Value = 28.13 P5 = Water In, Value = 30.53 P6 = Water Out, Value = 28.68 P7 = Ambient, Value = 24.03 P8 = Supply Pres(psi), Value = 123.35 P9 = Return Pres(psi), Value = 11.70 14:58:23 Magnet Current for MAIN, Imag = 164.9754 A Vtrans = 1.649808 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 14:58:24 MAIN (V) Voltage Measurement, V = 6.511246 V 14:58:27 Temperature Measurement, Probe Number = 0, T = 25.621000 C 14:58:28 Temperature Measurement, Probe Number = 1, T = 23.783000 C 14:58:28 Temperature Measurement, Probe Number = 2, T = 28.183000 C 14:58:29 Temperature Measurement, Probe Number = 3, T = 30.515000 C 14:58:30 Temperature Measurement, Probe Number = 4, T = 28.651000 C 14:58:30 Temperature Measurement, Probe Number = 5, T = 23.972000 C 14:58:31 Sensor Voltage Measurement, Sensor Number = 0, V = 2.646258 V 14:58:31 Sensor Measurement, Sensor Number = 0, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 2.646258 (V, A, ohm, ...), Measured Value = 123.469357 (C, atm, Hz, ...) 14:58:31 Sensor Voltage Measurement, Sensor Number = 1, V = 1.156062 V 14:58:31 Sensor Measurement, Sensor Number = 1, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 1.156062 (V, A, ohm, ...), Measured Value = 11.704620 (C, atm, Hz, ...) 14:58:31 Magnet Test Parameter Measurements P0 = MAIN, Value = 164.98 P1 = MAIN (V), Value = 6.51 P2 = Yoke, Value = 25.62 P3 = Bottom Coil, Value = 23.78 P4 = Top Coil, Value = 28.18 P5 = Water In, Value = 30.52 P6 = Water Out, Value = 28.65 P7 = Ambient, Value = 23.97 P8 = Supply Pres(psi), Value = 123.47 P9 = Return Pres(psi), Value = 11.70 14:59:32 Magnet Current for MAIN, Imag = 164.9753 A Vtrans = 1.649806 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 14:59:33 MAIN (V) Voltage Measurement, V = 6.511276 V 14:59:36 Temperature Measurement, Probe Number = 0, T = 25.628000 C 14:59:37 Temperature Measurement, Probe Number = 1, T = 23.846000 C 14:59:37 Temperature Measurement, Probe Number = 2, T = 28.298000 C 14:59:38 Temperature Measurement, Probe Number = 3, T = 30.505000 C 14:59:39 Temperature Measurement, Probe Number = 4, T = 28.639000 C 14:59:39 Temperature Measurement, Probe Number = 5, T = 23.994000 C 14:59:40 Sensor Voltage Measurement, Sensor Number = 0, V = 2.645199 V 14:59:40 Sensor Measurement, Sensor Number = 0, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 2.645199 (V, A, ohm, ...), Measured Value = 123.389940 (C, atm, Hz, ...) 14:59:40 Sensor Voltage Measurement, Sensor Number = 1, V = 1.156226 V 14:59:40 Sensor Measurement, Sensor Number = 1, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 1.156226 (V, A, ohm, ...), Measured Value = 11.716980 (C, atm, Hz, ...) 14:59:40 Magnet Test Parameter Measurements P0 = MAIN, Value = 164.98 P1 = MAIN (V), Value = 6.51 P2 = Yoke, Value = 25.63 P3 = Bottom Coil, Value = 23.85 P4 = Top Coil, Value = 28.30 P5 = Water In, Value = 30.50 P6 = Water Out, Value = 28.64 P7 = Ambient, Value = 23.99 P8 = Supply Pres(psi), Value = 123.39 P9 = Return Pres(psi), Value = 11.72 15:00:41 Magnet Current for MAIN, Imag = 164.9755 A Vtrans = 1.649809 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 15:00:42 MAIN (V) Voltage Measurement, V = 6.511286 V 15:00:45 Temperature Measurement, Probe Number = 0, T = 25.630000 C 15:00:46 Temperature Measurement, Probe Number = 1, T = 23.855000 C 15:00:46 Temperature Measurement, Probe Number = 2, T = 28.183000 C 15:00:47 Temperature Measurement, Probe Number = 3, T = 30.525000 C 15:00:48 Temperature Measurement, Probe Number = 4, T = 28.624000 C 15:00:48 Temperature Measurement, Probe Number = 5, T = 24.009000 C 15:00:49 Sensor Voltage Measurement, Sensor Number = 0, V = 2.645193 V 15:00:49 Sensor Measurement, Sensor Number = 0, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 2.645193 (V, A, ohm, ...), Measured Value = 123.389468 (C, atm, Hz, ...) 15:00:49 Sensor Voltage Measurement, Sensor Number = 1, V = 1.157819 V 15:00:49 Sensor Measurement, Sensor Number = 1, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 1.157819 (V, A, ohm, ...), Measured Value = 11.836388 (C, atm, Hz, ...) 15:00:49 Magnet Test Parameter Measurements P0 = MAIN, Value = 164.98 P1 = MAIN (V), Value = 6.51 P2 = Yoke, Value = 25.63 P3 = Bottom Coil, Value = 23.86 P4 = Top Coil, Value = 28.18 P5 = Water In, Value = 30.52 P6 = Water Out, Value = 28.62 P7 = Ambient, Value = 24.01 P8 = Supply Pres(psi), Value = 123.39 P9 = Return Pres(psi), Value = 11.84 15:01:50 Magnet Current for MAIN, Imag = 164.9755 A Vtrans = 1.649809 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 15:01:51 MAIN (V) Voltage Measurement, V = 6.511271 V 15:01:54 Temperature Measurement, Probe Number = 0, T = 25.648000 C 15:01:55 Temperature Measurement, Probe Number = 1, T = 23.897000 C 15:01:55 Temperature Measurement, Probe Number = 2, T = 28.126000 C 15:01:56 Temperature Measurement, Probe Number = 3, T = 30.521000 C 15:01:57 Temperature Measurement, Probe Number = 4, T = 28.606000 C 15:01:57 Temperature Measurement, Probe Number = 5, T = 24.065000 C 15:01:58 Sensor Voltage Measurement, Sensor Number = 0, V = 2.651063 V 15:01:58 Sensor Measurement, Sensor Number = 0, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 2.651063 (V, A, ohm, ...), Measured Value = 123.829717 (C, atm, Hz, ...) 15:01:58 Sensor Voltage Measurement, Sensor Number = 1, V = 1.154822 V 15:01:58 Sensor Measurement, Sensor Number = 1, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 1.154822 (V, A, ohm, ...), Measured Value = 11.611628 (C, atm, Hz, ...) 15:01:58 Magnet Test Parameter Measurements P0 = MAIN, Value = 164.98 P1 = MAIN (V), Value = 6.51 P2 = Yoke, Value = 25.65 P3 = Bottom Coil, Value = 23.90 P4 = Top Coil, Value = 28.13 P5 = Water In, Value = 30.52 P6 = Water Out, Value = 28.61 P7 = Ambient, Value = 24.07 P8 = Supply Pres(psi), Value = 123.83 P9 = Return Pres(psi), Value = 11.61 15:02:59 Magnet Current for MAIN, Imag = 164.9753 A Vtrans = 1.649806 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 15:03:00 MAIN (V) Voltage Measurement, V = 6.511439 V 15:03:03 Temperature Measurement, Probe Number = 0, T = 25.658000 C 15:03:04 Temperature Measurement, Probe Number = 1, T = 23.856000 C 15:03:04 Temperature Measurement, Probe Number = 2, T = 28.358000 C 15:03:05 Temperature Measurement, Probe Number = 3, T = 30.535000 C 15:03:06 Temperature Measurement, Probe Number = 4, T = 28.645000 C 15:03:06 Temperature Measurement, Probe Number = 5, T = 24.047000 C 15:03:07 Sensor Voltage Measurement, Sensor Number = 0, V = 2.630938 V 15:03:07 Sensor Measurement, Sensor Number = 0, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 2.630938 (V, A, ohm, ...), Measured Value = 122.320365 (C, atm, Hz, ...) 15:03:07 Sensor Voltage Measurement, Sensor Number = 1, V = 1.155667 V 15:03:07 Sensor Measurement, Sensor Number = 1, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 1.155667 (V, A, ohm, ...), Measured Value = 11.675025 (C, atm, Hz, ...) 15:03:07 Magnet Test Parameter Measurements P0 = MAIN, Value = 164.98 P1 = MAIN (V), Value = 6.51 P2 = Yoke, Value = 25.66 P3 = Bottom Coil, Value = 23.86 P4 = Top Coil, Value = 28.36 P5 = Water In, Value = 30.54 P6 = Water Out, Value = 28.64 P7 = Ambient, Value = 24.05 P8 = Supply Pres(psi), Value = 122.32 P9 = Return Pres(psi), Value = 11.68 15:04:08 Magnet Current for MAIN, Imag = 164.9752 A Vtrans = 1.649805 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 15:04:09 MAIN (V) Voltage Measurement, V = 6.511427 V 15:04:12 Temperature Measurement, Probe Number = 0, T = 25.665000 C 15:04:13 Temperature Measurement, Probe Number = 1, T = 23.830000 C 15:04:13 Temperature Measurement, Probe Number = 2, T = 28.513000 C 15:04:14 Temperature Measurement, Probe Number = 3, T = 30.540000 C 15:04:15 Temperature Measurement, Probe Number = 4, T = 28.613000 C 15:04:15 Temperature Measurement, Probe Number = 5, T = 24.125000 C 15:04:16 Sensor Voltage Measurement, Sensor Number = 0, V = 2.652083 V 15:04:16 Sensor Measurement, Sensor Number = 0, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 2.652083 (V, A, ohm, ...), Measured Value = 123.906233 (C, atm, Hz, ...) 15:04:16 Sensor Voltage Measurement, Sensor Number = 1, V = 1.157393 V 15:04:16 Sensor Measurement, Sensor Number = 1, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 1.157393 (V, A, ohm, ...), Measured Value = 11.804452 (C, atm, Hz, ...) 15:04:16 Magnet Test Parameter Measurements P0 = MAIN, Value = 164.98 P1 = MAIN (V), Value = 6.51 P2 = Yoke, Value = 25.66 P3 = Bottom Coil, Value = 23.83 P4 = Top Coil, Value = 28.51 P5 = Water In, Value = 30.54 P6 = Water Out, Value = 28.61 P7 = Ambient, Value = 24.12 P8 = Supply Pres(psi), Value = 123.91 P9 = Return Pres(psi), Value = 11.80 15:05:17 Magnet Current for MAIN, Imag = 164.9755 A Vtrans = 1.649809 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 15:05:18 MAIN (V) Voltage Measurement, V = 6.511307 V 15:05:21 Temperature Measurement, Probe Number = 0, T = 25.671000 C 15:05:22 Temperature Measurement, Probe Number = 1, T = 23.885000 C 15:05:22 Temperature Measurement, Probe Number = 2, T = 28.377000 C 15:05:23 Temperature Measurement, Probe Number = 3, T = 30.548000 C 15:05:24 Temperature Measurement, Probe Number = 4, T = 28.633000 C 15:05:24 Temperature Measurement, Probe Number = 5, T = 24.117000 C 15:05:25 Sensor Voltage Measurement, Sensor Number = 0, V = 2.652503 V 15:05:25 Sensor Measurement, Sensor Number = 0, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 2.652503 (V, A, ohm, ...), Measured Value = 123.937695 (C, atm, Hz, ...) 15:05:25 Sensor Voltage Measurement, Sensor Number = 1, V = 1.157878 V 15:05:25 Sensor Measurement, Sensor Number = 1, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 1.157878 (V, A, ohm, ...), Measured Value = 11.840880 (C, atm, Hz, ...) 15:05:25 Magnet Test Parameter Measurements P0 = MAIN, Value = 164.98 P1 = MAIN (V), Value = 6.51 P2 = Yoke, Value = 25.67 P3 = Bottom Coil, Value = 23.89 P4 = Top Coil, Value = 28.38 P5 = Water In, Value = 30.55 P6 = Water Out, Value = 28.63 P7 = Ambient, Value = 24.12 P8 = Supply Pres(psi), Value = 123.94 P9 = Return Pres(psi), Value = 11.84 15:06:26 Magnet Current for MAIN, Imag = 164.9758 A Vtrans = 1.649811 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 15:06:27 MAIN (V) Voltage Measurement, V = 6.511343 V 15:06:30 Temperature Measurement, Probe Number = 0, T = 25.686000 C 15:06:31 Temperature Measurement, Probe Number = 1, T = 23.939000 C 15:06:31 Temperature Measurement, Probe Number = 2, T = 28.218000 C 15:06:32 Temperature Measurement, Probe Number = 3, T = 30.536000 C 15:06:33 Temperature Measurement, Probe Number = 4, T = 28.648000 C 15:06:33 Temperature Measurement, Probe Number = 5, T = 24.168000 C 15:06:34 Sensor Voltage Measurement, Sensor Number = 0, V = 2.645072 V 15:06:34 Sensor Measurement, Sensor Number = 0, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 2.645072 (V, A, ohm, ...), Measured Value = 123.380385 (C, atm, Hz, ...) 15:06:34 Sensor Voltage Measurement, Sensor Number = 1, V = 1.155377 V 15:06:34 Sensor Measurement, Sensor Number = 1, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 1.155377 (V, A, ohm, ...), Measured Value = 11.653298 (C, atm, Hz, ...) 15:06:34 Magnet Test Parameter Measurements P0 = MAIN, Value = 164.98 P1 = MAIN (V), Value = 6.51 P2 = Yoke, Value = 25.69 P3 = Bottom Coil, Value = 23.94 P4 = Top Coil, Value = 28.22 P5 = Water In, Value = 30.54 P6 = Water Out, Value = 28.65 P7 = Ambient, Value = 24.17 P8 = Supply Pres(psi), Value = 123.38 P9 = Return Pres(psi), Value = 11.65 15:07:35 Magnet Current for MAIN, Imag = 164.9755 A Vtrans = 1.649809 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 15:07:36 MAIN (V) Voltage Measurement, V = 6.511496 V 15:07:39 Temperature Measurement, Probe Number = 0, T = 25.700000 C 15:07:40 Temperature Measurement, Probe Number = 1, T = 23.953000 C 15:07:40 Temperature Measurement, Probe Number = 2, T = 28.174000 C 15:07:41 Temperature Measurement, Probe Number = 3, T = 30.537000 C 15:07:42 Temperature Measurement, Probe Number = 4, T = 28.624000 C 15:07:42 Temperature Measurement, Probe Number = 5, T = 24.164000 C 15:07:43 Sensor Voltage Measurement, Sensor Number = 0, V = 2.666781 V 15:07:43 Sensor Measurement, Sensor Number = 0, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 2.666781 (V, A, ohm, ...), Measured Value = 125.008583 (C, atm, Hz, ...) 15:07:43 Sensor Voltage Measurement, Sensor Number = 1, V = 1.154484 V 15:07:43 Sensor Measurement, Sensor Number = 1, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 1.154484 (V, A, ohm, ...), Measured Value = 11.586337 (C, atm, Hz, ...) 15:07:43 Magnet Test Parameter Measurements P0 = MAIN, Value = 164.98 P1 = MAIN (V), Value = 6.51 P2 = Yoke, Value = 25.70 P3 = Bottom Coil, Value = 23.95 P4 = Top Coil, Value = 28.17 P5 = Water In, Value = 30.54 P6 = Water Out, Value = 28.62 P7 = Ambient, Value = 24.16 P8 = Supply Pres(psi), Value = 125.01 P9 = Return Pres(psi), Value = 11.59 15:08:44 Magnet Current for MAIN, Imag = 164.9758 A Vtrans = 1.649811 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 15:08:45 MAIN (V) Voltage Measurement, V = 6.511548 V 15:08:48 Temperature Measurement, Probe Number = 0, T = 25.713000 C 15:08:49 Temperature Measurement, Probe Number = 1, T = 23.950000 C 15:08:49 Temperature Measurement, Probe Number = 2, T = 28.084000 C 15:08:50 Temperature Measurement, Probe Number = 3, T = 30.553000 C 15:08:51 Temperature Measurement, Probe Number = 4, T = 28.664000 C 15:08:51 Temperature Measurement, Probe Number = 5, T = 24.154000 C 15:08:52 Sensor Voltage Measurement, Sensor Number = 0, V = 2.641770 V 15:08:52 Sensor Measurement, Sensor Number = 0, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 2.641770 (V, A, ohm, ...), Measured Value = 123.132773 (C, atm, Hz, ...) 15:08:52 Sensor Voltage Measurement, Sensor Number = 1, V = 1.155136 V 15:08:52 Sensor Measurement, Sensor Number = 1, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 1.155136 (V, A, ohm, ...), Measured Value = 11.635222 (C, atm, Hz, ...) 15:08:52 Magnet Test Parameter Measurements P0 = MAIN, Value = 164.98 P1 = MAIN (V), Value = 6.51 P2 = Yoke, Value = 25.71 P3 = Bottom Coil, Value = 23.95 P4 = Top Coil, Value = 28.08 P5 = Water In, Value = 30.55 P6 = Water Out, Value = 28.66 P7 = Ambient, Value = 24.15 P8 = Supply Pres(psi), Value = 123.13 P9 = Return Pres(psi), Value = 11.64 15:09:53 Magnet Current for MAIN, Imag = 164.9755 A Vtrans = 1.649809 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 15:09:54 MAIN (V) Voltage Measurement, V = 6.511662 V 15:09:57 Temperature Measurement, Probe Number = 0, T = 25.714000 C 15:09:58 Temperature Measurement, Probe Number = 1, T = 23.961000 C 15:09:59 Temperature Measurement, Probe Number = 2, T = 28.044000 C 15:09:59 Temperature Measurement, Probe Number = 3, T = 30.563000 C 15:10:00 Temperature Measurement, Probe Number = 4, T = 28.671000 C 15:10:00 Temperature Measurement, Probe Number = 5, T = 24.200000 C 15:10:01 Sensor Voltage Measurement, Sensor Number = 0, V = 2.653634 V 15:10:01 Sensor Measurement, Sensor Number = 0, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 2.653634 (V, A, ohm, ...), Measured Value = 124.022543 (C, atm, Hz, ...) 15:10:01 Sensor Voltage Measurement, Sensor Number = 1, V = 1.158289 V 15:10:01 Sensor Measurement, Sensor Number = 1, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 1.158289 (V, A, ohm, ...), Measured Value = 11.871690 (C, atm, Hz, ...) 15:10:01 Magnet Test Parameter Measurements P0 = MAIN, Value = 164.98 P1 = MAIN (V), Value = 6.51 P2 = Yoke, Value = 25.71 P3 = Bottom Coil, Value = 23.96 P4 = Top Coil, Value = 28.04 P5 = Water In, Value = 30.56 P6 = Water Out, Value = 28.67 P7 = Ambient, Value = 24.20 P8 = Supply Pres(psi), Value = 124.02 P9 = Return Pres(psi), Value = 11.87 15:11:02 Magnet Current for MAIN, Imag = 164.9757 A Vtrans = 1.649810 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 15:11:03 MAIN (V) Voltage Measurement, V = 6.511602 V 15:11:06 Temperature Measurement, Probe Number = 0, T = 25.720000 C 15:11:07 Temperature Measurement, Probe Number = 1, T = 23.939000 C 15:11:08 Temperature Measurement, Probe Number = 2, T = 28.062000 C 15:11:08 Temperature Measurement, Probe Number = 3, T = 30.562000 C 15:11:09 Temperature Measurement, Probe Number = 4, T = 28.635000 C 15:11:09 Temperature Measurement, Probe Number = 5, T = 24.180000 C 15:11:10 Sensor Voltage Measurement, Sensor Number = 0, V = 2.651325 V 15:11:10 Sensor Measurement, Sensor Number = 0, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 2.651325 (V, A, ohm, ...), Measured Value = 123.849383 (C, atm, Hz, ...) 15:11:10 Sensor Voltage Measurement, Sensor Number = 1, V = 1.154762 V 15:11:10 Sensor Measurement, Sensor Number = 1, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 1.154762 (V, A, ohm, ...), Measured Value = 11.607127 (C, atm, Hz, ...) 15:11:10 Magnet Test Parameter Measurements P0 = MAIN, Value = 164.98 P1 = MAIN (V), Value = 6.51 P2 = Yoke, Value = 25.72 P3 = Bottom Coil, Value = 23.94 P4 = Top Coil, Value = 28.06 P5 = Water In, Value = 30.56 P6 = Water Out, Value = 28.64 P7 = Ambient, Value = 24.18 P8 = Supply Pres(psi), Value = 123.85 P9 = Return Pres(psi), Value = 11.61 15:12:11 Magnet Current for MAIN, Imag = 164.9759 A Vtrans = 1.649813 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 15:12:12 MAIN (V) Voltage Measurement, V = 6.511533 V 15:12:15 Temperature Measurement, Probe Number = 0, T = 25.718000 C 15:12:16 Temperature Measurement, Probe Number = 1, T = 23.924000 C 15:12:17 Temperature Measurement, Probe Number = 2, T = 28.084000 C 15:12:17 Temperature Measurement, Probe Number = 3, T = 30.551000 C 15:12:18 Temperature Measurement, Probe Number = 4, T = 28.668000 C 15:12:18 Temperature Measurement, Probe Number = 5, T = 24.152000 C 15:12:19 Sensor Voltage Measurement, Sensor Number = 0, V = 2.650638 V 15:12:19 Sensor Measurement, Sensor Number = 0, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 2.650638 (V, A, ohm, ...), Measured Value = 123.797880 (C, atm, Hz, ...) 15:12:19 Sensor Voltage Measurement, Sensor Number = 1, V = 1.154466 V 15:12:19 Sensor Measurement, Sensor Number = 1, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 1.154466 (V, A, ohm, ...), Measured Value = 11.584935 (C, atm, Hz, ...) 15:12:19 Magnet Test Parameter Measurements P0 = MAIN, Value = 164.98 P1 = MAIN (V), Value = 6.51 P2 = Yoke, Value = 25.72 P3 = Bottom Coil, Value = 23.92 P4 = Top Coil, Value = 28.08 P5 = Water In, Value = 30.55 P6 = Water Out, Value = 28.67 P7 = Ambient, Value = 24.15 P8 = Supply Pres(psi), Value = 123.80 P9 = Return Pres(psi), Value = 11.58 15:13:20 Magnet Current for MAIN, Imag = 164.9754 A Vtrans = 1.649808 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 15:13:21 MAIN (V) Voltage Measurement, V = 6.511579 V 15:13:24 Temperature Measurement, Probe Number = 0, T = 25.727000 C 15:13:25 Temperature Measurement, Probe Number = 1, T = 23.940000 C 15:13:26 Temperature Measurement, Probe Number = 2, T = 28.089000 C 15:13:26 Temperature Measurement, Probe Number = 3, T = 30.549000 C 15:13:27 Temperature Measurement, Probe Number = 4, T = 28.677000 C 15:13:27 Temperature Measurement, Probe Number = 5, T = 24.170000 C 15:13:28 Sensor Voltage Measurement, Sensor Number = 0, V = 2.647617 V 15:13:28 Sensor Measurement, Sensor Number = 0, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 2.647617 (V, A, ohm, ...), Measured Value = 123.571245 (C, atm, Hz, ...) 15:13:28 Sensor Voltage Measurement, Sensor Number = 1, V = 1.156116 V 15:13:28 Sensor Measurement, Sensor Number = 1, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 1.156116 (V, A, ohm, ...), Measured Value = 11.708737 (C, atm, Hz, ...) 15:13:28 Magnet Test Parameter Measurements P0 = MAIN, Value = 164.98 P1 = MAIN (V), Value = 6.51 P2 = Yoke, Value = 25.73 P3 = Bottom Coil, Value = 23.94 P4 = Top Coil, Value = 28.09 P5 = Water In, Value = 30.55 P6 = Water Out, Value = 28.68 P7 = Ambient, Value = 24.17 P8 = Supply Pres(psi), Value = 123.57 P9 = Return Pres(psi), Value = 11.71 15:14:29 Magnet Current for MAIN, Imag = 164.9758 A Vtrans = 1.649811 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 15:14:30 MAIN (V) Voltage Measurement, V = 6.511628 V 15:14:33 Temperature Measurement, Probe Number = 0, T = 25.737000 C 15:14:34 Temperature Measurement, Probe Number = 1, T = 23.973000 C 15:14:35 Temperature Measurement, Probe Number = 2, T = 28.097000 C 15:14:35 Temperature Measurement, Probe Number = 3, T = 30.552000 C 15:14:36 Temperature Measurement, Probe Number = 4, T = 28.645000 C 15:14:36 Temperature Measurement, Probe Number = 5, T = 24.191000 C 15:14:37 Sensor Voltage Measurement, Sensor Number = 0, V = 2.651452 V 15:14:37 Sensor Measurement, Sensor Number = 0, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 2.651452 (V, A, ohm, ...), Measured Value = 123.858937 (C, atm, Hz, ...) 15:14:38 Sensor Voltage Measurement, Sensor Number = 1, V = 1.157645 V 15:14:38 Sensor Measurement, Sensor Number = 1, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 1.157645 (V, A, ohm, ...), Measured Value = 11.823367 (C, atm, Hz, ...) 15:14:38 Magnet Test Parameter Measurements P0 = MAIN, Value = 164.98 P1 = MAIN (V), Value = 6.51 P2 = Yoke, Value = 25.74 P3 = Bottom Coil, Value = 23.97 P4 = Top Coil, Value = 28.10 P5 = Water In, Value = 30.55 P6 = Water Out, Value = 28.64 P7 = Ambient, Value = 24.19 P8 = Supply Pres(psi), Value = 123.86 P9 = Return Pres(psi), Value = 11.82 15:15:38 Magnet Current for MAIN, Imag = 164.9760 A Vtrans = 1.649814 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 15:15:39 MAIN (V) Voltage Measurement, V = 6.511718 V 15:15:42 Temperature Measurement, Probe Number = 0, T = 25.742000 C 15:15:43 Temperature Measurement, Probe Number = 1, T = 23.978000 C 15:15:44 Temperature Measurement, Probe Number = 2, T = 28.049000 C 15:15:44 Temperature Measurement, Probe Number = 3, T = 30.564000 C 15:15:45 Temperature Measurement, Probe Number = 4, T = 28.642000 C 15:15:45 Temperature Measurement, Probe Number = 5, T = 24.233000 C 15:15:46 Sensor Voltage Measurement, Sensor Number = 0, V = 2.644188 V 15:15:46 Sensor Measurement, Sensor Number = 0, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 2.644188 (V, A, ohm, ...), Measured Value = 123.314085 (C, atm, Hz, ...) 15:15:47 Sensor Voltage Measurement, Sensor Number = 1, V = 1.154268 V 15:15:47 Sensor Measurement, Sensor Number = 1, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 1.154268 (V, A, ohm, ...), Measured Value = 11.570137 (C, atm, Hz, ...) 15:15:47 Magnet Test Parameter Measurements P0 = MAIN, Value = 164.98 P1 = MAIN (V), Value = 6.51 P2 = Yoke, Value = 25.74 P3 = Bottom Coil, Value = 23.98 P4 = Top Coil, Value = 28.05 P5 = Water In, Value = 30.56 P6 = Water Out, Value = 28.64 P7 = Ambient, Value = 24.23 P8 = Supply Pres(psi), Value = 123.31 P9 = Return Pres(psi), Value = 11.57 15:16:47 Magnet Current for MAIN, Imag = 164.9758 A Vtrans = 1.649811 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 15:16:48 MAIN (V) Voltage Measurement, V = 6.511817 V 15:16:51 Temperature Measurement, Probe Number = 0, T = 25.739000 C 15:16:52 Temperature Measurement, Probe Number = 1, T = 23.968000 C 15:16:53 Temperature Measurement, Probe Number = 2, T = 28.113000 C 15:16:53 Temperature Measurement, Probe Number = 3, T = 30.571000 C 15:16:54 Temperature Measurement, Probe Number = 4, T = 28.679000 C 15:16:54 Temperature Measurement, Probe Number = 5, T = 24.195000 C 15:16:55 Sensor Voltage Measurement, Sensor Number = 0, V = 2.664471 V 15:16:55 Sensor Measurement, Sensor Number = 0, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 2.664471 (V, A, ohm, ...), Measured Value = 124.835333 (C, atm, Hz, ...) 15:16:56 Sensor Voltage Measurement, Sensor Number = 1, V = 1.156284 V 15:16:56 Sensor Measurement, Sensor Number = 1, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 1.156284 (V, A, ohm, ...), Measured Value = 11.721293 (C, atm, Hz, ...) 15:16:56 Magnet Test Parameter Measurements P0 = MAIN, Value = 164.98 P1 = MAIN (V), Value = 6.51 P2 = Yoke, Value = 25.74 P3 = Bottom Coil, Value = 23.97 P4 = Top Coil, Value = 28.11 P5 = Water In, Value = 30.57 P6 = Water Out, Value = 28.68 P7 = Ambient, Value = 24.20 P8 = Supply Pres(psi), Value = 124.84 P9 = Return Pres(psi), Value = 11.72 15:17:56 Magnet Current for MAIN, Imag = 164.9759 A Vtrans = 1.649813 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 15:17:57 MAIN (V) Voltage Measurement, V = 6.511935 V 15:18:00 Temperature Measurement, Probe Number = 0, T = 25.747000 C 15:18:01 Temperature Measurement, Probe Number = 1, T = 23.977000 C 15:18:02 Temperature Measurement, Probe Number = 2, T = 28.055000 C 15:18:02 Temperature Measurement, Probe Number = 3, T = 30.570000 C 15:18:03 Temperature Measurement, Probe Number = 4, T = 28.703000 C 15:18:03 Temperature Measurement, Probe Number = 5, T = 24.226000 C 15:18:04 Sensor Voltage Measurement, Sensor Number = 0, V = 2.652115 V 15:18:04 Sensor Measurement, Sensor Number = 0, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 2.652115 (V, A, ohm, ...), Measured Value = 123.908662 (C, atm, Hz, ...) 15:18:05 Sensor Voltage Measurement, Sensor Number = 1, V = 1.155296 V 15:18:05 Sensor Measurement, Sensor Number = 1, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 1.155296 (V, A, ohm, ...), Measured Value = 11.647215 (C, atm, Hz, ...) 15:18:05 Magnet Test Parameter Measurements P0 = MAIN, Value = 164.98 P1 = MAIN (V), Value = 6.51 P2 = Yoke, Value = 25.75 P3 = Bottom Coil, Value = 23.98 P4 = Top Coil, Value = 28.05 P5 = Water In, Value = 30.57 P6 = Water Out, Value = 28.70 P7 = Ambient, Value = 24.23 P8 = Supply Pres(psi), Value = 123.91 P9 = Return Pres(psi), Value = 11.65 15:19:05 Magnet Current for MAIN, Imag = 164.9760 A Vtrans = 1.649814 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 15:19:06 MAIN (V) Voltage Measurement, V = 6.511909 V 15:19:09 Temperature Measurement, Probe Number = 0, T = 25.759000 C 15:19:10 Temperature Measurement, Probe Number = 1, T = 24.015000 C 15:19:11 Temperature Measurement, Probe Number = 2, T = 28.083000 C 15:19:11 Temperature Measurement, Probe Number = 3, T = 30.571000 C 15:19:12 Temperature Measurement, Probe Number = 4, T = 28.689000 C 15:19:12 Temperature Measurement, Probe Number = 5, T = 24.234000 C 15:19:13 Sensor Voltage Measurement, Sensor Number = 0, V = 2.646683 V 15:19:13 Sensor Measurement, Sensor Number = 0, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 2.646683 (V, A, ohm, ...), Measured Value = 123.501195 (C, atm, Hz, ...) 15:19:14 Sensor Voltage Measurement, Sensor Number = 1, V = 1.157066 V 15:19:14 Sensor Measurement, Sensor Number = 1, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 1.157066 (V, A, ohm, ...), Measured Value = 11.779913 (C, atm, Hz, ...) 15:19:14 Magnet Test Parameter Measurements P0 = MAIN, Value = 164.98 P1 = MAIN (V), Value = 6.51 P2 = Yoke, Value = 25.76 P3 = Bottom Coil, Value = 24.02 P4 = Top Coil, Value = 28.08 P5 = Water In, Value = 30.57 P6 = Water Out, Value = 28.69 P7 = Ambient, Value = 24.23 P8 = Supply Pres(psi), Value = 123.50 P9 = Return Pres(psi), Value = 11.78 15:20:14 Magnet Current for MAIN, Imag = 164.9758 A Vtrans = 1.649811 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 15:20:15 MAIN (V) Voltage Measurement, V = 6.511938 V 15:20:18 Temperature Measurement, Probe Number = 0, T = 25.775000 C 15:20:19 Temperature Measurement, Probe Number = 1, T = 23.997000 C 15:20:20 Temperature Measurement, Probe Number = 2, T = 28.048000 C 15:20:20 Temperature Measurement, Probe Number = 3, T = 30.573000 C 15:20:21 Temperature Measurement, Probe Number = 4, T = 28.700000 C 15:20:21 Temperature Measurement, Probe Number = 5, T = 24.241000 C 15:20:22 Sensor Voltage Measurement, Sensor Number = 0, V = 2.651284 V 15:20:22 Sensor Measurement, Sensor Number = 0, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 2.651284 (V, A, ohm, ...), Measured Value = 123.846293 (C, atm, Hz, ...) 15:20:23 Sensor Voltage Measurement, Sensor Number = 1, V = 1.157518 V 15:20:23 Sensor Measurement, Sensor Number = 1, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 1.157518 (V, A, ohm, ...), Measured Value = 11.813820 (C, atm, Hz, ...) 15:20:23 Magnet Test Parameter Measurements P0 = MAIN, Value = 164.98 P1 = MAIN (V), Value = 6.51 P2 = Yoke, Value = 25.77 P3 = Bottom Coil, Value = 24.00 P4 = Top Coil, Value = 28.05 P5 = Water In, Value = 30.57 P6 = Water Out, Value = 28.70 P7 = Ambient, Value = 24.24 P8 = Supply Pres(psi), Value = 123.85 P9 = Return Pres(psi), Value = 11.81 15:21:23 Magnet Current for MAIN, Imag = 164.9764 A Vtrans = 1.649818 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 15:21:24 MAIN (V) Voltage Measurement, V = 6.511813 V 15:21:27 Temperature Measurement, Probe Number = 0, T = 25.778000 C 15:21:28 Temperature Measurement, Probe Number = 1, T = 23.956000 C 15:21:29 Temperature Measurement, Probe Number = 2, T = 28.045000 C 15:21:29 Temperature Measurement, Probe Number = 3, T = 30.569000 C 15:21:30 Temperature Measurement, Probe Number = 4, T = 28.710000 C 15:21:30 Temperature Measurement, Probe Number = 5, T = 24.239000 C 15:21:31 Sensor Voltage Measurement, Sensor Number = 0, V = 2.633857 V 15:21:31 Sensor Measurement, Sensor Number = 0, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 2.633857 (V, A, ohm, ...), Measured Value = 122.539312 (C, atm, Hz, ...) 15:21:32 Sensor Voltage Measurement, Sensor Number = 1, V = 1.157575 V 15:21:32 Sensor Measurement, Sensor Number = 1, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 1.157575 (V, A, ohm, ...), Measured Value = 11.818125 (C, atm, Hz, ...) 15:21:32 Magnet Test Parameter Measurements P0 = MAIN, Value = 164.98 P1 = MAIN (V), Value = 6.51 P2 = Yoke, Value = 25.78 P3 = Bottom Coil, Value = 23.96 P4 = Top Coil, Value = 28.05 P5 = Water In, Value = 30.57 P6 = Water Out, Value = 28.71 P7 = Ambient, Value = 24.24 P8 = Supply Pres(psi), Value = 122.54 P9 = Return Pres(psi), Value = 11.82 15:22:32 Magnet Current for MAIN, Imag = 164.9758 A Vtrans = 1.649811 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 15:22:33 MAIN (V) Voltage Measurement, V = 6.511809 V 15:22:37 Temperature Measurement, Probe Number = 0, T = 25.783000 C 15:22:37 Temperature Measurement, Probe Number = 1, T = 23.930000 C 15:22:38 Temperature Measurement, Probe Number = 2, T = 28.065000 C 15:22:38 Temperature Measurement, Probe Number = 3, T = 30.574000 C 15:22:39 Temperature Measurement, Probe Number = 4, T = 28.717000 C 15:22:39 Temperature Measurement, Probe Number = 5, T = 24.268000 C 15:22:40 Sensor Voltage Measurement, Sensor Number = 0, V = 2.660554 V 15:22:40 Sensor Measurement, Sensor Number = 0, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 2.660554 (V, A, ohm, ...), Measured Value = 124.541550 (C, atm, Hz, ...) 15:22:41 Sensor Voltage Measurement, Sensor Number = 1, V = 1.156746 V 15:22:41 Sensor Measurement, Sensor Number = 1, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 1.156746 (V, A, ohm, ...), Measured Value = 11.755942 (C, atm, Hz, ...) 15:22:41 Magnet Test Parameter Measurements P0 = MAIN, Value = 164.98 P1 = MAIN (V), Value = 6.51 P2 = Yoke, Value = 25.78 P3 = Bottom Coil, Value = 23.93 P4 = Top Coil, Value = 28.07 P5 = Water In, Value = 30.57 P6 = Water Out, Value = 28.72 P7 = Ambient, Value = 24.27 P8 = Supply Pres(psi), Value = 124.54 P9 = Return Pres(psi), Value = 11.76 15:23:41 Magnet Current for MAIN, Imag = 164.9760 A Vtrans = 1.649814 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 15:23:42 MAIN (V) Voltage Measurement, V = 6.511672 V 15:23:46 Temperature Measurement, Probe Number = 0, T = 25.790000 C 15:23:46 Temperature Measurement, Probe Number = 1, T = 24.005000 C 15:23:47 Temperature Measurement, Probe Number = 2, T = 28.089000 C 15:23:47 Temperature Measurement, Probe Number = 3, T = 30.572000 C 15:23:48 Temperature Measurement, Probe Number = 4, T = 28.706000 C 15:23:48 Temperature Measurement, Probe Number = 5, T = 24.296000 C 15:23:49 Sensor Voltage Measurement, Sensor Number = 0, V = 2.654595 V 15:23:49 Sensor Measurement, Sensor Number = 0, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 2.654595 (V, A, ohm, ...), Measured Value = 124.094655 (C, atm, Hz, ...) 15:23:50 Sensor Voltage Measurement, Sensor Number = 1, V = 1.155220 V 15:23:50 Sensor Measurement, Sensor Number = 1, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 1.155220 (V, A, ohm, ...), Measured Value = 11.641500 (C, atm, Hz, ...) 15:23:50 Magnet Test Parameter Measurements P0 = MAIN, Value = 164.98 P1 = MAIN (V), Value = 6.51 P2 = Yoke, Value = 25.79 P3 = Bottom Coil, Value = 24.00 P4 = Top Coil, Value = 28.09 P5 = Water In, Value = 30.57 P6 = Water Out, Value = 28.71 P7 = Ambient, Value = 24.30 P8 = Supply Pres(psi), Value = 124.09 P9 = Return Pres(psi), Value = 11.64 15:24:50 Magnet Current for MAIN, Imag = 164.9763 A Vtrans = 1.649816 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 15:24:51 MAIN (V) Voltage Measurement, V = 6.511389 V 15:24:55 Temperature Measurement, Probe Number = 0, T = 25.797000 C 15:24:55 Temperature Measurement, Probe Number = 1, T = 24.062000 C 15:24:56 Temperature Measurement, Probe Number = 2, T = 28.109000 C 15:24:56 Temperature Measurement, Probe Number = 3, T = 30.573000 C 15:24:57 Temperature Measurement, Probe Number = 4, T = 28.692000 C 15:24:58 Temperature Measurement, Probe Number = 5, T = 24.241000 C 15:24:58 Sensor Voltage Measurement, Sensor Number = 0, V = 2.659804 V 15:24:58 Sensor Measurement, Sensor Number = 0, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 2.659804 (V, A, ohm, ...), Measured Value = 124.485270 (C, atm, Hz, ...) 15:24:59 Sensor Voltage Measurement, Sensor Number = 1, V = 1.155673 V 15:24:59 Sensor Measurement, Sensor Number = 1, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 1.155673 (V, A, ohm, ...), Measured Value = 11.675497 (C, atm, Hz, ...) 15:24:59 Magnet Test Parameter Measurements P0 = MAIN, Value = 164.98 P1 = MAIN (V), Value = 6.51 P2 = Yoke, Value = 25.80 P3 = Bottom Coil, Value = 24.06 P4 = Top Coil, Value = 28.11 P5 = Water In, Value = 30.57 P6 = Water Out, Value = 28.69 P7 = Ambient, Value = 24.24 P8 = Supply Pres(psi), Value = 124.49 P9 = Return Pres(psi), Value = 11.68 15:25:59 Magnet Current for MAIN, Imag = 164.9760 A Vtrans = 1.649814 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 15:26:00 MAIN (V) Voltage Measurement, V = 6.511084 V 15:26:04 Temperature Measurement, Probe Number = 0, T = 25.804000 C 15:26:04 Temperature Measurement, Probe Number = 1, T = 24.004000 C 15:26:05 Temperature Measurement, Probe Number = 2, T = 28.127000 C 15:26:05 Temperature Measurement, Probe Number = 3, T = 30.552000 C 15:26:06 Temperature Measurement, Probe Number = 4, T = 28.687000 C 15:26:07 Temperature Measurement, Probe Number = 5, T = 24.269000 C 15:26:07 Sensor Voltage Measurement, Sensor Number = 0, V = 2.653786 V 15:26:07 Sensor Measurement, Sensor Number = 0, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 2.653786 (V, A, ohm, ...), Measured Value = 124.033973 (C, atm, Hz, ...) 15:26:08 Sensor Voltage Measurement, Sensor Number = 1, V = 1.156007 V 15:26:08 Sensor Measurement, Sensor Number = 1, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 1.156007 (V, A, ohm, ...), Measured Value = 11.700503 (C, atm, Hz, ...) 15:26:08 Magnet Test Parameter Measurements P0 = MAIN, Value = 164.98 P1 = MAIN (V), Value = 6.51 P2 = Yoke, Value = 25.80 P3 = Bottom Coil, Value = 24.00 P4 = Top Coil, Value = 28.13 P5 = Water In, Value = 30.55 P6 = Water Out, Value = 28.69 P7 = Ambient, Value = 24.27 P8 = Supply Pres(psi), Value = 124.03 P9 = Return Pres(psi), Value = 11.70 15:27:08 Magnet Current for MAIN, Imag = 164.9762 A Vtrans = 1.649815 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 15:27:09 MAIN (V) Voltage Measurement, V = 6.510688 V 15:27:13 Temperature Measurement, Probe Number = 0, T = 25.813000 C 15:27:13 Temperature Measurement, Probe Number = 1, T = 24.022000 C 15:27:14 Temperature Measurement, Probe Number = 2, T = 28.193000 C 15:27:14 Temperature Measurement, Probe Number = 3, T = 30.556000 C 15:27:15 Temperature Measurement, Probe Number = 4, T = 28.666000 C 15:27:16 Temperature Measurement, Probe Number = 5, T = 24.265000 C 15:27:16 Sensor Voltage Measurement, Sensor Number = 0, V = 2.632885 V 15:27:16 Sensor Measurement, Sensor Number = 0, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 2.632885 (V, A, ohm, ...), Measured Value = 122.466360 (C, atm, Hz, ...) 15:27:17 Sensor Voltage Measurement, Sensor Number = 1, V = 1.156666 V 15:27:17 Sensor Measurement, Sensor Number = 1, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 1.156666 (V, A, ohm, ...), Measured Value = 11.749950 (C, atm, Hz, ...) 15:27:17 Magnet Test Parameter Measurements P0 = MAIN, Value = 164.98 P1 = MAIN (V), Value = 6.51 P2 = Yoke, Value = 25.81 P3 = Bottom Coil, Value = 24.02 P4 = Top Coil, Value = 28.19 P5 = Water In, Value = 30.56 P6 = Water Out, Value = 28.67 P7 = Ambient, Value = 24.27 P8 = Supply Pres(psi), Value = 122.47 P9 = Return Pres(psi), Value = 11.75 15:28:17 Magnet Current for MAIN, Imag = 164.9765 A Vtrans = 1.649819 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 15:28:18 MAIN (V) Voltage Measurement, V = 6.510474 V 15:28:22 Temperature Measurement, Probe Number = 0, T = 25.824000 C 15:28:22 Temperature Measurement, Probe Number = 1, T = 24.045000 C 15:28:23 Temperature Measurement, Probe Number = 2, T = 28.119000 C 15:28:23 Temperature Measurement, Probe Number = 3, T = 30.541000 C 15:28:24 Temperature Measurement, Probe Number = 4, T = 28.647000 C 15:28:25 Temperature Measurement, Probe Number = 5, T = 24.315000 C 15:28:25 Sensor Voltage Measurement, Sensor Number = 0, V = 2.660176 V 15:28:25 Sensor Measurement, Sensor Number = 0, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 2.660176 (V, A, ohm, ...), Measured Value = 124.513177 (C, atm, Hz, ...) 15:28:26 Sensor Voltage Measurement, Sensor Number = 1, V = 1.154935 V 15:28:26 Sensor Measurement, Sensor Number = 1, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 1.154935 (V, A, ohm, ...), Measured Value = 11.620147 (C, atm, Hz, ...) 15:28:26 Magnet Test Parameter Measurements P0 = MAIN, Value = 164.98 P1 = MAIN (V), Value = 6.51 P2 = Yoke, Value = 25.82 P3 = Bottom Coil, Value = 24.05 P4 = Top Coil, Value = 28.12 P5 = Water In, Value = 30.54 P6 = Water Out, Value = 28.65 P7 = Ambient, Value = 24.32 P8 = Supply Pres(psi), Value = 124.51 P9 = Return Pres(psi), Value = 11.62 15:29:26 Magnet Current for MAIN, Imag = 164.9763 A Vtrans = 1.649816 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 15:29:27 MAIN (V) Voltage Measurement, V = 6.510670 V 15:29:31 Temperature Measurement, Probe Number = 0, T = 25.835000 C 15:29:31 Temperature Measurement, Probe Number = 1, T = 24.063000 C 15:29:32 Temperature Measurement, Probe Number = 2, T = 28.095000 C 15:29:32 Temperature Measurement, Probe Number = 3, T = 30.541000 C 15:29:33 Temperature Measurement, Probe Number = 4, T = 28.628000 C 15:29:34 Temperature Measurement, Probe Number = 5, T = 24.333000 C 15:29:34 Sensor Voltage Measurement, Sensor Number = 0, V = 2.657405 V 15:29:34 Sensor Measurement, Sensor Number = 0, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 2.657405 (V, A, ohm, ...), Measured Value = 124.305368 (C, atm, Hz, ...) 15:29:35 Sensor Voltage Measurement, Sensor Number = 1, V = 1.155078 V 15:29:35 Sensor Measurement, Sensor Number = 1, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 1.155078 (V, A, ohm, ...), Measured Value = 11.630828 (C, atm, Hz, ...) 15:29:35 Magnet Test Parameter Measurements P0 = MAIN, Value = 164.98 P1 = MAIN (V), Value = 6.51 P2 = Yoke, Value = 25.84 P3 = Bottom Coil, Value = 24.06 P4 = Top Coil, Value = 28.09 P5 = Water In, Value = 30.54 P6 = Water Out, Value = 28.63 P7 = Ambient, Value = 24.33 P8 = Supply Pres(psi), Value = 124.31 P9 = Return Pres(psi), Value = 11.63 15:30:35 Magnet Current for MAIN, Imag = 164.9767 A Vtrans = 1.649820 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 15:30:36 MAIN (V) Voltage Measurement, V = 6.510774 V 15:30:40 Temperature Measurement, Probe Number = 0, T = 25.838000 C 15:30:40 Temperature Measurement, Probe Number = 1, T = 24.052000 C 15:30:41 Temperature Measurement, Probe Number = 2, T = 28.140000 C 15:30:41 Temperature Measurement, Probe Number = 3, T = 30.537000 C 15:30:42 Temperature Measurement, Probe Number = 4, T = 28.659000 C 15:30:43 Temperature Measurement, Probe Number = 5, T = 24.287000 C 15:30:43 Sensor Voltage Measurement, Sensor Number = 0, V = 2.652833 V 15:30:43 Sensor Measurement, Sensor Number = 0, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 2.652833 (V, A, ohm, ...), Measured Value = 123.962512 (C, atm, Hz, ...) 15:30:44 Sensor Voltage Measurement, Sensor Number = 1, V = 1.154643 V 15:30:44 Sensor Measurement, Sensor Number = 1, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 1.154643 (V, A, ohm, ...), Measured Value = 11.598232 (C, atm, Hz, ...) 15:30:44 Magnet Test Parameter Measurements P0 = MAIN, Value = 164.98 P1 = MAIN (V), Value = 6.51 P2 = Yoke, Value = 25.84 P3 = Bottom Coil, Value = 24.05 P4 = Top Coil, Value = 28.14 P5 = Water In, Value = 30.54 P6 = Water Out, Value = 28.66 P7 = Ambient, Value = 24.29 P8 = Supply Pres(psi), Value = 123.96 P9 = Return Pres(psi), Value = 11.60 15:31:44 Magnet Current for MAIN, Imag = 164.9762 A Vtrans = 1.649815 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 15:31:45 MAIN (V) Voltage Measurement, V = 6.511005 V 15:31:49 Temperature Measurement, Probe Number = 0, T = 25.848000 C 15:31:49 Temperature Measurement, Probe Number = 1, T = 24.080000 C 15:31:50 Temperature Measurement, Probe Number = 2, T = 28.096000 C 15:31:50 Temperature Measurement, Probe Number = 3, T = 30.557000 C 15:31:51 Temperature Measurement, Probe Number = 4, T = 28.698000 C 15:31:52 Temperature Measurement, Probe Number = 5, T = 24.334000 C 15:31:52 Sensor Voltage Measurement, Sensor Number = 0, V = 2.649664 V 15:31:52 Sensor Measurement, Sensor Number = 0, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 2.649664 (V, A, ohm, ...), Measured Value = 123.724830 (C, atm, Hz, ...) 15:31:53 Sensor Voltage Measurement, Sensor Number = 1, V = 1.155789 V 15:31:53 Sensor Measurement, Sensor Number = 1, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 1.155789 (V, A, ohm, ...), Measured Value = 11.684205 (C, atm, Hz, ...) 15:31:53 Magnet Test Parameter Measurements P0 = MAIN, Value = 164.98 P1 = MAIN (V), Value = 6.51 P2 = Yoke, Value = 25.85 P3 = Bottom Coil, Value = 24.08 P4 = Top Coil, Value = 28.10 P5 = Water In, Value = 30.56 P6 = Water Out, Value = 28.70 P7 = Ambient, Value = 24.33 P8 = Supply Pres(psi), Value = 123.72 P9 = Return Pres(psi), Value = 11.68 15:32:53 Magnet Current for MAIN, Imag = 164.9758 A Vtrans = 1.649811 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 15:32:54 MAIN (V) Voltage Measurement, V = 6.511283 V 15:32:58 Temperature Measurement, Probe Number = 0, T = 25.855000 C 15:32:58 Temperature Measurement, Probe Number = 1, T = 24.054000 C 15:32:59 Temperature Measurement, Probe Number = 2, T = 28.103000 C 15:32:59 Temperature Measurement, Probe Number = 3, T = 30.565000 C 15:33:00 Temperature Measurement, Probe Number = 4, T = 28.702000 C 15:33:01 Temperature Measurement, Probe Number = 5, T = 24.323000 C 15:33:01 Sensor Voltage Measurement, Sensor Number = 0, V = 2.653251 V 15:33:01 Sensor Measurement, Sensor Number = 0, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 2.653251 (V, A, ohm, ...), Measured Value = 123.993795 (C, atm, Hz, ...) 15:33:02 Sensor Voltage Measurement, Sensor Number = 1, V = 1.156525 V 15:33:02 Sensor Measurement, Sensor Number = 1, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 1.156525 (V, A, ohm, ...), Measured Value = 11.739367 (C, atm, Hz, ...) 15:33:02 Magnet Test Parameter Measurements P0 = MAIN, Value = 164.98 P1 = MAIN (V), Value = 6.51 P2 = Yoke, Value = 25.86 P3 = Bottom Coil, Value = 24.05 P4 = Top Coil, Value = 28.10 P5 = Water In, Value = 30.57 P6 = Water Out, Value = 28.70 P7 = Ambient, Value = 24.32 P8 = Supply Pres(psi), Value = 123.99 P9 = Return Pres(psi), Value = 11.74 15:34:02 Magnet Current for MAIN, Imag = 164.9764 A Vtrans = 1.649818 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 15:34:03 MAIN (V) Voltage Measurement, V = 6.511534 V 15:34:07 Temperature Measurement, Probe Number = 0, T = 25.860000 C 15:34:07 Temperature Measurement, Probe Number = 1, T = 24.078000 C 15:34:08 Temperature Measurement, Probe Number = 2, T = 28.106000 C 15:34:08 Temperature Measurement, Probe Number = 3, T = 30.555000 C 15:34:09 Temperature Measurement, Probe Number = 4, T = 28.666000 C 15:34:10 Temperature Measurement, Probe Number = 5, T = 24.345000 C 15:34:10 Sensor Voltage Measurement, Sensor Number = 0, V = 2.649653 V 15:34:10 Sensor Measurement, Sensor Number = 0, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 2.649653 (V, A, ohm, ...), Measured Value = 123.723990 (C, atm, Hz, ...) 15:34:11 Sensor Voltage Measurement, Sensor Number = 1, V = 1.156052 V 15:34:11 Sensor Measurement, Sensor Number = 1, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 1.156052 (V, A, ohm, ...), Measured Value = 11.703870 (C, atm, Hz, ...) 15:34:11 Magnet Test Parameter Measurements P0 = MAIN, Value = 164.98 P1 = MAIN (V), Value = 6.51 P2 = Yoke, Value = 25.86 P3 = Bottom Coil, Value = 24.08 P4 = Top Coil, Value = 28.11 P5 = Water In, Value = 30.55 P6 = Water Out, Value = 28.67 P7 = Ambient, Value = 24.34 P8 = Supply Pres(psi), Value = 123.72 P9 = Return Pres(psi), Value = 11.70 15:35:12 Magnet Current for MAIN, Imag = 164.9765 A Vtrans = 1.649819 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 15:35:12 MAIN (V) Voltage Measurement, V = 6.511679 V 15:35:16 Temperature Measurement, Probe Number = 0, T = 25.865000 C 15:35:16 Temperature Measurement, Probe Number = 1, T = 24.068000 C 15:35:17 Temperature Measurement, Probe Number = 2, T = 28.154000 C 15:35:17 Temperature Measurement, Probe Number = 3, T = 30.573000 C 15:35:18 Temperature Measurement, Probe Number = 4, T = 28.719000 C 15:35:19 Temperature Measurement, Probe Number = 5, T = 24.357000 C 15:35:19 Sensor Voltage Measurement, Sensor Number = 0, V = 2.656234 V 15:35:19 Sensor Measurement, Sensor Number = 0, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 2.656234 (V, A, ohm, ...), Measured Value = 124.217520 (C, atm, Hz, ...) 15:35:20 Sensor Voltage Measurement, Sensor Number = 1, V = 1.157800 V 15:35:20 Sensor Measurement, Sensor Number = 1, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 1.157800 (V, A, ohm, ...), Measured Value = 11.834977 (C, atm, Hz, ...) 15:35:20 Magnet Test Parameter Measurements P0 = MAIN, Value = 164.98 P1 = MAIN (V), Value = 6.51 P2 = Yoke, Value = 25.86 P3 = Bottom Coil, Value = 24.07 P4 = Top Coil, Value = 28.15 P5 = Water In, Value = 30.57 P6 = Water Out, Value = 28.72 P7 = Ambient, Value = 24.36 P8 = Supply Pres(psi), Value = 124.22 P9 = Return Pres(psi), Value = 11.83 15:36:21 Magnet Current for MAIN, Imag = 164.9760 A Vtrans = 1.649814 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 15:36:21 MAIN (V) Voltage Measurement, V = 6.511845 V 15:36:25 Temperature Measurement, Probe Number = 0, T = 25.876000 C 15:36:25 Temperature Measurement, Probe Number = 1, T = 24.119000 C 15:36:26 Temperature Measurement, Probe Number = 2, T = 28.132000 C 15:36:26 Temperature Measurement, Probe Number = 3, T = 30.590000 C 15:36:27 Temperature Measurement, Probe Number = 4, T = 28.720000 C 15:36:28 Temperature Measurement, Probe Number = 5, T = 24.351000 C 15:36:28 Sensor Voltage Measurement, Sensor Number = 0, V = 2.655421 V 15:36:28 Sensor Measurement, Sensor Number = 0, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 2.655421 (V, A, ohm, ...), Measured Value = 124.156560 (C, atm, Hz, ...) 15:36:29 Sensor Voltage Measurement, Sensor Number = 1, V = 1.157092 V 15:36:29 Sensor Measurement, Sensor Number = 1, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 1.157092 (V, A, ohm, ...), Measured Value = 11.781885 (C, atm, Hz, ...) 15:36:29 Magnet Test Parameter Measurements P0 = MAIN, Value = 164.98 P1 = MAIN (V), Value = 6.51 P2 = Yoke, Value = 25.88 P3 = Bottom Coil, Value = 24.12 P4 = Top Coil, Value = 28.13 P5 = Water In, Value = 30.59 P6 = Water Out, Value = 28.72 P7 = Ambient, Value = 24.35 P8 = Supply Pres(psi), Value = 124.16 P9 = Return Pres(psi), Value = 11.78 15:37:30 Magnet Current for MAIN, Imag = 164.9765 A Vtrans = 1.649819 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 15:37:30 MAIN (V) Voltage Measurement, V = 6.512013 V 15:37:34 Temperature Measurement, Probe Number = 0, T = 25.886000 C 15:37:34 Temperature Measurement, Probe Number = 1, T = 24.119000 C 15:37:35 Temperature Measurement, Probe Number = 2, T = 28.300000 C 15:37:35 Temperature Measurement, Probe Number = 3, T = 30.579000 C 15:37:36 Temperature Measurement, Probe Number = 4, T = 28.641000 C 15:37:37 Temperature Measurement, Probe Number = 5, T = 24.381000 C 15:37:37 Sensor Voltage Measurement, Sensor Number = 0, V = 2.655670 V 15:37:37 Sensor Measurement, Sensor Number = 0, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 2.655670 (V, A, ohm, ...), Measured Value = 124.175287 (C, atm, Hz, ...) 15:37:38 Sensor Voltage Measurement, Sensor Number = 1, V = 1.157152 V 15:37:38 Sensor Measurement, Sensor Number = 1, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 1.157152 (V, A, ohm, ...), Measured Value = 11.786378 (C, atm, Hz, ...) 15:37:38 Magnet Test Parameter Measurements P0 = MAIN, Value = 164.98 P1 = MAIN (V), Value = 6.51 P2 = Yoke, Value = 25.89 P3 = Bottom Coil, Value = 24.12 P4 = Top Coil, Value = 28.30 P5 = Water In, Value = 30.58 P6 = Water Out, Value = 28.64 P7 = Ambient, Value = 24.38 P8 = Supply Pres(psi), Value = 124.18 P9 = Return Pres(psi), Value = 11.79 15:38:39 Magnet Current for MAIN, Imag = 164.9762 A Vtrans = 1.649815 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 15:38:39 MAIN (V) Voltage Measurement, V = 6.512050 V 15:38:43 Temperature Measurement, Probe Number = 0, T = 25.890000 C 15:38:43 Temperature Measurement, Probe Number = 1, T = 24.128000 C 15:38:44 Temperature Measurement, Probe Number = 2, T = 28.446000 C 15:38:44 Temperature Measurement, Probe Number = 3, T = 30.586000 C 15:38:45 Temperature Measurement, Probe Number = 4, T = 28.678000 C 15:38:46 Temperature Measurement, Probe Number = 5, T = 24.391000 C 15:38:46 Sensor Voltage Measurement, Sensor Number = 0, V = 2.649034 V 15:38:46 Sensor Measurement, Sensor Number = 0, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 2.649034 (V, A, ohm, ...), Measured Value = 123.677535 (C, atm, Hz, ...) 15:38:47 Sensor Voltage Measurement, Sensor Number = 1, V = 1.154859 V 15:38:47 Sensor Measurement, Sensor Number = 1, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 1.154859 (V, A, ohm, ...), Measured Value = 11.614432 (C, atm, Hz, ...) 15:38:47 Magnet Test Parameter Measurements P0 = MAIN, Value = 164.98 P1 = MAIN (V), Value = 6.51 P2 = Yoke, Value = 25.89 P3 = Bottom Coil, Value = 24.13 P4 = Top Coil, Value = 28.45 P5 = Water In, Value = 30.59 P6 = Water Out, Value = 28.68 P7 = Ambient, Value = 24.39 P8 = Supply Pres(psi), Value = 123.68 P9 = Return Pres(psi), Value = 11.61 15:39:48 Magnet Current for MAIN, Imag = 164.9764 A Vtrans = 1.649818 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 15:39:48 MAIN (V) Voltage Measurement, V = 6.512272 V 15:39:52 Temperature Measurement, Probe Number = 0, T = 25.895000 C 15:39:52 Temperature Measurement, Probe Number = 1, T = 24.108000 C 15:39:53 Temperature Measurement, Probe Number = 2, T = 28.415000 C 15:39:54 Temperature Measurement, Probe Number = 3, T = 30.595000 C 15:39:54 Temperature Measurement, Probe Number = 4, T = 28.702000 C 15:39:55 Temperature Measurement, Probe Number = 5, T = 24.396000 C 15:39:55 Sensor Voltage Measurement, Sensor Number = 0, V = 2.653169 V 15:39:55 Sensor Measurement, Sensor Number = 0, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 2.653169 (V, A, ohm, ...), Measured Value = 123.987705 (C, atm, Hz, ...) 15:39:56 Sensor Voltage Measurement, Sensor Number = 1, V = 1.157666 V 15:39:56 Sensor Measurement, Sensor Number = 1, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 1.157666 (V, A, ohm, ...), Measured Value = 11.824957 (C, atm, Hz, ...) 15:39:56 Magnet Test Parameter Measurements P0 = MAIN, Value = 164.98 P1 = MAIN (V), Value = 6.51 P2 = Yoke, Value = 25.89 P3 = Bottom Coil, Value = 24.11 P4 = Top Coil, Value = 28.41 P5 = Water In, Value = 30.59 P6 = Water Out, Value = 28.70 P7 = Ambient, Value = 24.40 P8 = Supply Pres(psi), Value = 123.99 P9 = Return Pres(psi), Value = 11.82 15:40:57 Magnet Current for MAIN, Imag = 164.9765 A Vtrans = 1.649819 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 15:40:57 MAIN (V) Voltage Measurement, V = 6.512194 V 15:41:01 Temperature Measurement, Probe Number = 0, T = 25.899000 C 15:41:01 Temperature Measurement, Probe Number = 1, T = 24.114000 C 15:41:02 Temperature Measurement, Probe Number = 2, T = 28.286000 C 15:41:03 Temperature Measurement, Probe Number = 3, T = 30.594000 C 15:41:03 Temperature Measurement, Probe Number = 4, T = 28.724000 C 15:41:04 Temperature Measurement, Probe Number = 5, T = 24.392000 C 15:41:04 Sensor Voltage Measurement, Sensor Number = 0, V = 2.645925 V 15:41:04 Sensor Measurement, Sensor Number = 0, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 2.645925 (V, A, ohm, ...), Measured Value = 123.444353 (C, atm, Hz, ...) 15:41:05 Sensor Voltage Measurement, Sensor Number = 1, V = 1.156185 V 15:41:05 Sensor Measurement, Sensor Number = 1, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 1.156185 (V, A, ohm, ...), Measured Value = 11.713890 (C, atm, Hz, ...) 15:41:05 Magnet Test Parameter Measurements P0 = MAIN, Value = 164.98 P1 = MAIN (V), Value = 6.51 P2 = Yoke, Value = 25.90 P3 = Bottom Coil, Value = 24.11 P4 = Top Coil, Value = 28.29 P5 = Water In, Value = 30.59 P6 = Water Out, Value = 28.72 P7 = Ambient, Value = 24.39 P8 = Supply Pres(psi), Value = 123.44 P9 = Return Pres(psi), Value = 11.71 15:42:06 Magnet Current for MAIN, Imag = 164.9767 A Vtrans = 1.649820 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 15:42:06 MAIN (V) Voltage Measurement, V = 6.512277 V 15:42:10 Temperature Measurement, Probe Number = 0, T = 25.900000 C 15:42:10 Temperature Measurement, Probe Number = 1, T = 24.140000 C 15:42:11 Temperature Measurement, Probe Number = 2, T = 28.288000 C 15:42:12 Temperature Measurement, Probe Number = 3, T = 30.603000 C 15:42:12 Temperature Measurement, Probe Number = 4, T = 28.696000 C 15:42:13 Temperature Measurement, Probe Number = 5, T = 24.362000 C 15:42:13 Sensor Voltage Measurement, Sensor Number = 0, V = 2.657268 V 15:42:13 Sensor Measurement, Sensor Number = 0, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 2.657268 (V, A, ohm, ...), Measured Value = 124.295063 (C, atm, Hz, ...) 15:42:14 Sensor Voltage Measurement, Sensor Number = 1, V = 1.155878 V 15:42:14 Sensor Measurement, Sensor Number = 1, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 1.155878 (V, A, ohm, ...), Measured Value = 11.690858 (C, atm, Hz, ...) 15:42:14 Magnet Test Parameter Measurements P0 = MAIN, Value = 164.98 P1 = MAIN (V), Value = 6.51 P2 = Yoke, Value = 25.90 P3 = Bottom Coil, Value = 24.14 P4 = Top Coil, Value = 28.29 P5 = Water In, Value = 30.60 P6 = Water Out, Value = 28.70 P7 = Ambient, Value = 24.36 P8 = Supply Pres(psi), Value = 124.30 P9 = Return Pres(psi), Value = 11.69 15:43:15 Magnet Current for MAIN, Imag = 164.9764 A Vtrans = 1.649818 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 15:43:15 MAIN (V) Voltage Measurement, V = 6.512373 V 15:43:19 Temperature Measurement, Probe Number = 0, T = 25.898000 C 15:43:19 Temperature Measurement, Probe Number = 1, T = 24.127000 C 15:43:20 Temperature Measurement, Probe Number = 2, T = 28.282000 C 15:43:21 Temperature Measurement, Probe Number = 3, T = 30.607000 C 15:43:21 Temperature Measurement, Probe Number = 4, T = 28.697000 C 15:43:22 Temperature Measurement, Probe Number = 5, T = 24.379000 C 15:43:22 Sensor Voltage Measurement, Sensor Number = 0, V = 2.642072 V 15:43:22 Sensor Measurement, Sensor Number = 0, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 2.642072 (V, A, ohm, ...), Measured Value = 123.155437 (C, atm, Hz, ...) 15:43:23 Sensor Voltage Measurement, Sensor Number = 1, V = 1.155206 V 15:43:23 Sensor Measurement, Sensor Number = 1, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 1.155206 (V, A, ohm, ...), Measured Value = 11.640472 (C, atm, Hz, ...) 15:43:23 Magnet Test Parameter Measurements P0 = MAIN, Value = 164.98 P1 = MAIN (V), Value = 6.51 P2 = Yoke, Value = 25.90 P3 = Bottom Coil, Value = 24.13 P4 = Top Coil, Value = 28.28 P5 = Water In, Value = 30.61 P6 = Water Out, Value = 28.70 P7 = Ambient, Value = 24.38 P8 = Supply Pres(psi), Value = 123.16 P9 = Return Pres(psi), Value = 11.64 15:44:24 Magnet Current for MAIN, Imag = 164.9767 A Vtrans = 1.649820 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 15:44:24 MAIN (V) Voltage Measurement, V = 6.512269 V 15:44:28 Temperature Measurement, Probe Number = 0, T = 25.912000 C 15:44:28 Temperature Measurement, Probe Number = 1, T = 24.129000 C 15:44:29 Temperature Measurement, Probe Number = 2, T = 28.268000 C 15:44:30 Temperature Measurement, Probe Number = 3, T = 30.610000 C 15:44:30 Temperature Measurement, Probe Number = 4, T = 28.709000 C 15:44:31 Temperature Measurement, Probe Number = 5, T = 24.382000 C 15:44:31 Sensor Voltage Measurement, Sensor Number = 0, V = 2.643552 V 15:44:31 Sensor Measurement, Sensor Number = 0, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 2.643552 (V, A, ohm, ...), Measured Value = 123.266415 (C, atm, Hz, ...) 15:44:32 Sensor Voltage Measurement, Sensor Number = 1, V = 1.155692 V 15:44:32 Sensor Measurement, Sensor Number = 1, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 1.155692 (V, A, ohm, ...), Measured Value = 11.676900 (C, atm, Hz, ...) 15:44:32 Magnet Test Parameter Measurements P0 = MAIN, Value = 164.98 P1 = MAIN (V), Value = 6.51 P2 = Yoke, Value = 25.91 P3 = Bottom Coil, Value = 24.13 P4 = Top Coil, Value = 28.27 P5 = Water In, Value = 30.61 P6 = Water Out, Value = 28.71 P7 = Ambient, Value = 24.38 P8 = Supply Pres(psi), Value = 123.27 P9 = Return Pres(psi), Value = 11.68 15:45:33 Magnet Current for MAIN, Imag = 164.9769 A Vtrans = 1.649823 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 15:45:33 MAIN (V) Voltage Measurement, V = 6.512388 V 15:45:37 Temperature Measurement, Probe Number = 0, T = 25.927000 C 15:45:37 Temperature Measurement, Probe Number = 1, T = 24.140000 C 15:45:38 Temperature Measurement, Probe Number = 2, T = 28.255000 C 15:45:39 Temperature Measurement, Probe Number = 3, T = 30.617000 C 15:45:39 Temperature Measurement, Probe Number = 4, T = 28.728000 C 15:45:40 Temperature Measurement, Probe Number = 5, T = 24.429000 C 15:45:40 Sensor Voltage Measurement, Sensor Number = 0, V = 2.666654 V 15:45:40 Sensor Measurement, Sensor Number = 0, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 2.666654 (V, A, ohm, ...), Measured Value = 124.999035 (C, atm, Hz, ...) 15:45:41 Sensor Voltage Measurement, Sensor Number = 1, V = 1.157101 V 15:45:41 Sensor Measurement, Sensor Number = 1, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 1.157101 (V, A, ohm, ...), Measured Value = 11.782538 (C, atm, Hz, ...) 15:45:41 Magnet Test Parameter Measurements P0 = MAIN, Value = 164.98 P1 = MAIN (V), Value = 6.51 P2 = Yoke, Value = 25.93 P3 = Bottom Coil, Value = 24.14 P4 = Top Coil, Value = 28.25 P5 = Water In, Value = 30.62 P6 = Water Out, Value = 28.73 P7 = Ambient, Value = 24.43 P8 = Supply Pres(psi), Value = 125.00 P9 = Return Pres(psi), Value = 11.78 15:46:42 Magnet Current for MAIN, Imag = 164.9768 A Vtrans = 1.649821 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 15:46:42 MAIN (V) Voltage Measurement, V = 6.512296 V 15:46:46 Temperature Measurement, Probe Number = 0, T = 25.919000 C 15:46:46 Temperature Measurement, Probe Number = 1, T = 24.192000 C 15:46:47 Temperature Measurement, Probe Number = 2, T = 28.371000 C 15:46:48 Temperature Measurement, Probe Number = 3, T = 30.613000 C 15:46:48 Temperature Measurement, Probe Number = 4, T = 28.668000 C 15:46:49 Temperature Measurement, Probe Number = 5, T = 24.425000 C 15:46:49 Sensor Voltage Measurement, Sensor Number = 0, V = 2.648840 V 15:46:49 Sensor Measurement, Sensor Number = 0, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 2.648840 (V, A, ohm, ...), Measured Value = 123.663022 (C, atm, Hz, ...) 15:46:50 Sensor Voltage Measurement, Sensor Number = 1, V = 1.154232 V 15:46:50 Sensor Measurement, Sensor Number = 1, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 1.154232 (V, A, ohm, ...), Measured Value = 11.567423 (C, atm, Hz, ...) 15:46:50 Magnet Test Parameter Measurements P0 = MAIN, Value = 164.98 P1 = MAIN (V), Value = 6.51 P2 = Yoke, Value = 25.92 P3 = Bottom Coil, Value = 24.19 P4 = Top Coil, Value = 28.37 P5 = Water In, Value = 30.61 P6 = Water Out, Value = 28.67 P7 = Ambient, Value = 24.43 P8 = Supply Pres(psi), Value = 123.66 P9 = Return Pres(psi), Value = 11.57 15:47:51 Magnet Current for MAIN, Imag = 164.9770 A Vtrans = 1.649824 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 15:47:51 MAIN (V) Voltage Measurement, V = 6.512210 V 15:47:55 Temperature Measurement, Probe Number = 0, T = 25.940000 C 15:47:55 Temperature Measurement, Probe Number = 1, T = 24.223000 C 15:47:56 Temperature Measurement, Probe Number = 2, T = 28.468000 C 15:47:57 Temperature Measurement, Probe Number = 3, T = 30.601000 C 15:47:57 Temperature Measurement, Probe Number = 4, T = 28.657000 C 15:47:58 Temperature Measurement, Probe Number = 5, T = 24.425000 C 15:47:58 Sensor Voltage Measurement, Sensor Number = 0, V = 2.651499 V 15:47:58 Sensor Measurement, Sensor Number = 0, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 2.651499 (V, A, ohm, ...), Measured Value = 123.862402 (C, atm, Hz, ...) 15:47:59 Sensor Voltage Measurement, Sensor Number = 1, V = 1.155744 V 15:47:59 Sensor Measurement, Sensor Number = 1, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 1.155744 (V, A, ohm, ...), Measured Value = 11.680830 (C, atm, Hz, ...) 15:47:59 Magnet Test Parameter Measurements P0 = MAIN, Value = 164.98 P1 = MAIN (V), Value = 6.51 P2 = Yoke, Value = 25.94 P3 = Bottom Coil, Value = 24.22 P4 = Top Coil, Value = 28.47 P5 = Water In, Value = 30.60 P6 = Water Out, Value = 28.66 P7 = Ambient, Value = 24.43 P8 = Supply Pres(psi), Value = 123.86 P9 = Return Pres(psi), Value = 11.68 15:49:00 Magnet Current for MAIN, Imag = 164.9768 A Vtrans = 1.649821 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 15:49:00 MAIN (V) Voltage Measurement, V = 6.512202 V 15:49:04 Temperature Measurement, Probe Number = 0, T = 25.954000 C 15:49:04 Temperature Measurement, Probe Number = 1, T = 24.215000 C 15:49:05 Temperature Measurement, Probe Number = 2, T = 28.578000 C 15:49:06 Temperature Measurement, Probe Number = 3, T = 30.593000 C 15:49:06 Temperature Measurement, Probe Number = 4, T = 28.689000 C 15:49:07 Temperature Measurement, Probe Number = 5, T = 24.405000 C 15:49:07 Sensor Voltage Measurement, Sensor Number = 0, V = 2.649041 V 15:49:07 Sensor Measurement, Sensor Number = 0, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 2.649041 (V, A, ohm, ...), Measured Value = 123.678097 (C, atm, Hz, ...) 15:49:08 Sensor Voltage Measurement, Sensor Number = 1, V = 1.156092 V 15:49:08 Sensor Measurement, Sensor Number = 1, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 1.156092 (V, A, ohm, ...), Measured Value = 11.706870 (C, atm, Hz, ...) 15:49:08 Magnet Test Parameter Measurements P0 = MAIN, Value = 164.98 P1 = MAIN (V), Value = 6.51 P2 = Yoke, Value = 25.95 P3 = Bottom Coil, Value = 24.21 P4 = Top Coil, Value = 28.58 P5 = Water In, Value = 30.59 P6 = Water Out, Value = 28.69 P7 = Ambient, Value = 24.41 P8 = Supply Pres(psi), Value = 123.68 P9 = Return Pres(psi), Value = 11.71 15:50:09 Magnet Current for MAIN, Imag = 164.9773 A Vtrans = 1.649826 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 15:50:09 MAIN (V) Voltage Measurement, V = 6.512360 V 15:50:13 Temperature Measurement, Probe Number = 0, T = 25.959000 C 15:50:13 Temperature Measurement, Probe Number = 1, T = 24.200000 C 15:50:14 Temperature Measurement, Probe Number = 2, T = 28.426000 C 15:50:15 Temperature Measurement, Probe Number = 3, T = 30.604000 C 15:50:15 Temperature Measurement, Probe Number = 4, T = 28.692000 C 15:50:16 Temperature Measurement, Probe Number = 5, T = 24.449000 C 15:50:16 Sensor Voltage Measurement, Sensor Number = 0, V = 2.643713 V 15:50:16 Sensor Measurement, Sensor Number = 0, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 2.643713 (V, A, ohm, ...), Measured Value = 123.278498 (C, atm, Hz, ...) 15:50:17 Sensor Voltage Measurement, Sensor Number = 1, V = 1.154870 V 15:50:17 Sensor Measurement, Sensor Number = 1, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 1.154870 (V, A, ohm, ...), Measured Value = 11.615280 (C, atm, Hz, ...) 15:50:17 Magnet Test Parameter Measurements P0 = MAIN, Value = 164.98 P1 = MAIN (V), Value = 6.51 P2 = Yoke, Value = 25.96 P3 = Bottom Coil, Value = 24.20 P4 = Top Coil, Value = 28.43 P5 = Water In, Value = 30.60 P6 = Water Out, Value = 28.69 P7 = Ambient, Value = 24.45 P8 = Supply Pres(psi), Value = 123.28 P9 = Return Pres(psi), Value = 11.62 15:51:18 Magnet Current for MAIN, Imag = 164.9767 A Vtrans = 1.649820 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 15:51:18 MAIN (V) Voltage Measurement, V = 6.512241 V 15:51:22 Temperature Measurement, Probe Number = 0, T = 25.963000 C 15:51:22 Temperature Measurement, Probe Number = 1, T = 24.207000 C 15:51:23 Temperature Measurement, Probe Number = 2, T = 28.351000 C 15:51:24 Temperature Measurement, Probe Number = 3, T = 30.615000 C 15:51:24 Temperature Measurement, Probe Number = 4, T = 28.675000 C 15:51:25 Temperature Measurement, Probe Number = 5, T = 24.444000 C 15:51:25 Sensor Voltage Measurement, Sensor Number = 0, V = 2.652164 V 15:51:25 Sensor Measurement, Sensor Number = 0, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 2.652164 (V, A, ohm, ...), Measured Value = 123.912315 (C, atm, Hz, ...) 15:51:26 Sensor Voltage Measurement, Sensor Number = 1, V = 1.155633 V 15:51:26 Sensor Measurement, Sensor Number = 1, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 1.155633 (V, A, ohm, ...), Measured Value = 11.672498 (C, atm, Hz, ...) 15:51:26 Magnet Test Parameter Measurements P0 = MAIN, Value = 164.98 P1 = MAIN (V), Value = 6.51 P2 = Yoke, Value = 25.96 P3 = Bottom Coil, Value = 24.21 P4 = Top Coil, Value = 28.35 P5 = Water In, Value = 30.61 P6 = Water Out, Value = 28.68 P7 = Ambient, Value = 24.44 P8 = Supply Pres(psi), Value = 123.91 P9 = Return Pres(psi), Value = 11.67 15:52:27 Magnet Current for MAIN, Imag = 164.9769 A Vtrans = 1.649823 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 15:52:27 MAIN (V) Voltage Measurement, V = 6.512234 V 15:52:31 Temperature Measurement, Probe Number = 0, T = 25.969000 C 15:52:31 Temperature Measurement, Probe Number = 1, T = 24.186000 C 15:52:32 Temperature Measurement, Probe Number = 2, T = 28.238000 C 15:52:33 Temperature Measurement, Probe Number = 3, T = 30.612000 C 15:52:33 Temperature Measurement, Probe Number = 4, T = 28.715000 C 15:52:34 Temperature Measurement, Probe Number = 5, T = 24.445000 C 15:52:34 Sensor Voltage Measurement, Sensor Number = 0, V = 2.676880 V 15:52:34 Sensor Measurement, Sensor Number = 0, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 2.676880 (V, A, ohm, ...), Measured Value = 125.766023 (C, atm, Hz, ...) 15:52:35 Sensor Voltage Measurement, Sensor Number = 1, V = 1.156067 V 15:52:35 Sensor Measurement, Sensor Number = 1, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 1.156067 (V, A, ohm, ...), Measured Value = 11.704995 (C, atm, Hz, ...) 15:52:35 Magnet Test Parameter Measurements P0 = MAIN, Value = 164.98 P1 = MAIN (V), Value = 6.51 P2 = Yoke, Value = 25.97 P3 = Bottom Coil, Value = 24.19 P4 = Top Coil, Value = 28.24 P5 = Water In, Value = 30.61 P6 = Water Out, Value = 28.71 P7 = Ambient, Value = 24.45 P8 = Supply Pres(psi), Value = 125.77 P9 = Return Pres(psi), Value = 11.70 15:52:35 Magnet system MAIN current ramping to 0.0000 A... 15:52:56 Magnet system MAIN current ramping to 185.0000 A... 15:53:19 Magnet system MAIN current ramping to 0.0000 A... 15:53:42 Magnet system MAIN current ramping to 185.0000 A... 15:54:04 Magnet system MAIN current ramping to 0.0000 A... 15:54:27 Magnet system MAIN current ramping to 185.0000 A... 15:54:50 Magnet system MAIN current ramping to 0.0000 A... 15:55:15 Position moved: 0.000000 m 15:55:18 Magnet system MAIN current ramping to 0.0000 A... 15:55:29 Magnet Current for MAIN, Imag = 0.0169 A Vtrans = 0.000223 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 15:55:30 MAIN (V) Voltage Measurement, V = 0.000702 V 15:55:33 Temperature Measurement, Probe Number = 0, T = 25.985000 C 15:55:34 Temperature Measurement, Probe Number = 1, T = 24.171000 C 15:55:34 Temperature Measurement, Probe Number = 2, T = 27.763000 C 15:55:35 Temperature Measurement, Probe Number = 3, T = 29.725000 C 15:55:36 Temperature Measurement, Probe Number = 4, T = 28.716000 C 15:55:36 Temperature Measurement, Probe Number = 5, T = 24.435000 C 15:55:37 Sensor Voltage Measurement, Sensor Number = 0, V = 2.640438 V 15:55:37 Sensor Measurement, Sensor Number = 0, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 2.640438 (V, A, ohm, ...), Measured Value = 123.032850 (C, atm, Hz, ...) 15:55:37 Sensor Voltage Measurement, Sensor Number = 1, V = 1.154401 V 15:55:37 Sensor Measurement, Sensor Number = 1, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 1.154401 (V, A, ohm, ...), Measured Value = 11.580067 (C, atm, Hz, ...) 15:55:37 Magnet Test Parameter Measurements P0 = MAIN, Value = 0.02 P1 = MAIN (V), Value = 0.00 P2 = Yoke, Value = 25.98 P3 = Bottom Coil, Value = 24.17 P4 = Top Coil, Value = 27.76 P5 = Water In, Value = 29.73 P6 = Water Out, Value = 28.72 P7 = Ambient, Value = 24.43 P8 = Supply Pres(psi), Value = 123.03 P9 = Return Pres(psi), Value = 11.58 15:55:37 Magnetic Field Measurement Measuring the magnet strength... 15:55:37 Integrated Voltage Measurement... 15:55:38 Magnet Current for MAIN, Imag = 0.0169 A Vtrans = 0.000222 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 15:55:47 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = -0.000002718 Vs 15:55:54 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.000016419 Vs 15:55:55 Magnet Current for MAIN, Imag = 0.0174 A Vtrans = 0.000227 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 15:56:03 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = -0.000002556 Vs 15:56:10 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.000016297 Vs 15:56:11 Magnet Current for MAIN, Imag = 0.0174 A Vtrans = 0.000227 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 15:56:19 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = -0.000002680 Vs 15:56:26 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.000016023 Vs 15:56:27 Magnet Current for MAIN, Imag = 0.0174 A Vtrans = 0.000228 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 15:56:35 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = -0.000001510 Vs 15:56:42 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.000015251 Vs 15:56:42 Integrated Voltage Average and RMS of Measurements X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.000006816 +- 0.000000084 Vs 15:56:42 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0000000 m, dx = 0.0100000 m, Nturns = 10.0 VT = 0.000006816 +- 0.000000084 Vs BL = 0.00006816 +- 0.00000084 Tm 15:56:46 Position moved: 0.000000 m 15:56:46 BL vs X Fit Coefficients: Cn sigCn n (Tm/m^n) ----- --------+-------- 0 0.00007 0.00000 Compute Fit Residuals: X0 BL sigBL BLfit BL-BLfit (m) (Tm) (Tm) (Tm) (Tm) ----------- --------+-------- -------- -------- 0.0000000 0.00007 0.00000 0.00007 0.00000 15:56:46 Magnet Strength Calculation Harmonic Number = 1 Number Of X0 Positions = 1 X0 BL sigBL (m) (Tm) (Tm) -------- --------+-------- 0.0000 0.00007 0.00000 SL = 0.0000682 +- 0.0000008 Tm^2/m^n 15:56:46 Magnet Strength Measurement Harmonic Number = 1 SL = 0.0000682 +- 0.0000008 Tm^2/m^n 15:56:49 Position moved: 0.000000 m 15:56:49 Magnet system MAIN current ramping to 15.0000 A... 15:57:01 Magnet Current for MAIN, Imag = 14.9973 A Vtrans = 0.150026 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 15:57:02 MAIN (V) Voltage Measurement, V = 0.588998 V 15:57:05 Temperature Measurement, Probe Number = 0, T = 25.992000 C 15:57:06 Temperature Measurement, Probe Number = 1, T = 24.099000 C 15:57:07 Temperature Measurement, Probe Number = 2, T = 27.468000 C 15:57:07 Temperature Measurement, Probe Number = 3, T = 28.974000 C 15:57:08 Temperature Measurement, Probe Number = 4, T = 28.697000 C 15:57:08 Temperature Measurement, Probe Number = 5, T = 24.435000 C 15:57:09 Sensor Voltage Measurement, Sensor Number = 0, V = 2.646103 V 15:57:09 Sensor Measurement, Sensor Number = 0, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 2.646103 (V, A, ohm, ...), Measured Value = 123.457740 (C, atm, Hz, ...) 15:57:10 Sensor Voltage Measurement, Sensor Number = 1, V = 1.156891 V 15:57:10 Sensor Measurement, Sensor Number = 1, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 1.156891 (V, A, ohm, ...), Measured Value = 11.766802 (C, atm, Hz, ...) 15:57:10 Magnet Test Parameter Measurements P0 = MAIN, Value = 15.00 P1 = MAIN (V), Value = 0.59 P2 = Yoke, Value = 25.99 P3 = Bottom Coil, Value = 24.10 P4 = Top Coil, Value = 27.47 P5 = Water In, Value = 28.97 P6 = Water Out, Value = 28.70 P7 = Ambient, Value = 24.43 P8 = Supply Pres(psi), Value = 123.46 P9 = Return Pres(psi), Value = 11.77 15:57:10 Magnetic Field Measurement Measuring the magnet strength... 15:57:10 Integrated Voltage Measurement... 15:57:10 Magnet Current for MAIN, Imag = 14.9975 A Vtrans = 0.150029 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 15:57:19 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.000570304 Vs 15:57:27 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.000588646 Vs 15:57:27 Magnet Current for MAIN, Imag = 14.9976 A Vtrans = 0.150030 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 15:57:35 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.000569782 Vs 15:57:43 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.000588132 Vs 15:57:43 Magnet Current for MAIN, Imag = 14.9975 A Vtrans = 0.150028 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 15:57:51 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.000569471 Vs 15:57:59 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.000588927 Vs 15:57:59 Magnet Current for MAIN, Imag = 14.9972 A Vtrans = 0.150025 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 15:58:07 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.000570048 Vs 15:58:15 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.000589477 Vs 15:58:15 Integrated Voltage Average and RMS of Measurements X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.000579349 +- 0.000000301 Vs 15:58:15 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0000000 m, dx = 0.0100000 m, Nturns = 10.0 VT = 0.000579349 +- 0.000000301 Vs BL = 0.00579349 +- 0.00000301 Tm 15:58:18 Position moved: 0.000000 m 15:58:18 BL vs X Fit Coefficients: Cn sigCn n (Tm/m^n) ----- --------+-------- 0 0.00579 0.00000 Compute Fit Residuals: X0 BL sigBL BLfit BL-BLfit (m) (Tm) (Tm) (Tm) (Tm) ----------- --------+-------- -------- -------- 0.0000000 0.00579 0.00000 0.00579 0.00000 15:58:18 Magnet Strength Calculation Harmonic Number = 1 Number Of X0 Positions = 1 X0 BL sigBL (m) (Tm) (Tm) -------- --------+-------- 0.0000 0.00579 0.00000 SL = 0.0057935 +- 0.0000030 Tm^2/m^n 15:58:18 Magnet Strength Measurement Harmonic Number = 1 SL = 0.0057935 +- 0.0000030 Tm^2/m^n Remeasure 15:58:18 Magnetic Field Measurement Measuring the magnet strength... 15:58:18 Integrated Voltage Measurement... 15:58:19 Magnet Current for MAIN, Imag = 14.9974 A Vtrans = 0.150028 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 15:58:28 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.000568842 Vs 15:58:35 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.000587335 Vs 15:58:36 Magnet Current for MAIN, Imag = 14.9975 A Vtrans = 0.150029 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 15:58:44 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.000569650 Vs 15:58:51 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.000588038 Vs 15:58:52 Magnet Current for MAIN, Imag = 14.9975 A Vtrans = 0.150029 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 15:59:00 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.000570599 Vs 15:59:07 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.000587956 Vs 15:59:08 Magnet Current for MAIN, Imag = 14.9976 A Vtrans = 0.150029 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 15:59:16 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.000570415 Vs 15:59:23 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.000587751 Vs 15:59:23 Integrated Voltage Average and RMS of Measurements X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.000578823 +- 0.000000451 Vs 15:59:23 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0000000 m, dx = 0.0100000 m, Nturns = 10.0 VT = 0.000578823 +- 0.000000451 Vs BL = 0.00578823 +- 0.00000451 Tm 15:59:27 Position moved: 0.000000 m 15:59:27 BL vs X Fit Coefficients: Cn sigCn n (Tm/m^n) ----- --------+-------- 0 0.00579 0.00000 Compute Fit Residuals: X0 BL sigBL BLfit BL-BLfit (m) (Tm) (Tm) (Tm) (Tm) ----------- --------+-------- -------- -------- 0.0000000 0.00579 0.00000 0.00579 0.00000 15:59:27 Magnet Strength Calculation Harmonic Number = 1 Number Of X0 Positions = 1 X0 BL sigBL (m) (Tm) (Tm) -------- --------+-------- 0.0000 0.00579 0.00000 SL = 0.0057882 +- 0.0000045 Tm^2/m^n 15:59:27 Magnet Strength Measurement Harmonic Number = 1 SL = 0.0057882 +- 0.0000045 Tm^2/m^n Remeasure 15:59:27 Magnetic Field Measurement Measuring the magnet strength... 15:59:27 Integrated Voltage Measurement... 15:59:28 Magnet Current for MAIN, Imag = 14.9977 A Vtrans = 0.150030 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 15:59:36 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.000570286 Vs 15:59:44 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.000587231 Vs 15:59:45 Magnet Current for MAIN, Imag = 14.9977 A Vtrans = 0.150030 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 15:59:52 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.000571149 Vs 16:00:00 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.000586386 Vs 16:00:01 Magnet Current for MAIN, Imag = 14.9978 A Vtrans = 0.150031 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:00:08 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.000571727 Vs 16:00:16 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.000586660 Vs 16:00:16 Magnet Current for MAIN, Imag = 14.9976 A Vtrans = 0.150030 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:00:24 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.000572032 Vs 16:00:32 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.000586520 Vs 16:00:32 Integrated Voltage Average and RMS of Measurements X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.000578999 +- 0.000000238 Vs 16:00:32 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0000000 m, dx = 0.0100000 m, Nturns = 10.0 VT = 0.000578999 +- 0.000000238 Vs BL = 0.00578999 +- 0.00000238 Tm 16:00:35 Position moved: 0.000000 m 16:00:35 BL vs X Fit Coefficients: Cn sigCn n (Tm/m^n) ----- --------+-------- 0 0.00579 0.00000 Compute Fit Residuals: X0 BL sigBL BLfit BL-BLfit (m) (Tm) (Tm) (Tm) (Tm) ----------- --------+-------- -------- -------- 0.0000000 0.00579 0.00000 0.00579 0.00000 16:00:35 Magnet Strength Calculation Harmonic Number = 1 Number Of X0 Positions = 1 X0 BL sigBL (m) (Tm) (Tm) -------- --------+-------- 0.0000 0.00579 0.00000 SL = 0.0057900 +- 0.0000024 Tm^2/m^n 16:00:35 Magnet Strength Measurement Harmonic Number = 1 SL = 0.0057900 +- 0.0000024 Tm^2/m^n Remeasure 16:00:35 Magnetic Field Measurement Measuring the magnet strength... 16:00:35 Integrated Voltage Measurement... 16:00:36 Magnet Current for MAIN, Imag = 14.9976 A Vtrans = 0.150029 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:00:45 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.000572163 Vs 16:00:53 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.000585462 Vs 16:00:53 Magnet Current for MAIN, Imag = 14.9975 A Vtrans = 0.150029 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:01:01 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.000572356 Vs 16:01:09 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.000585463 Vs 16:01:09 Magnet Current for MAIN, Imag = 14.9978 A Vtrans = 0.150031 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:01:17 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.000572024 Vs 16:01:25 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.000585889 Vs 16:01:25 Magnet Current for MAIN, Imag = 14.9976 A Vtrans = 0.150029 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:01:33 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.000570803 Vs 16:01:41 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.000587962 Vs 16:01:41 Integrated Voltage Average and RMS of Measurements X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.000579015 +- 0.000000218 Vs 16:01:41 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0000000 m, dx = 0.0100000 m, Nturns = 10.0 VT = 0.000579015 +- 0.000000218 Vs BL = 0.00579015 +- 0.00000218 Tm 16:01:44 Position moved: 0.000000 m 16:01:44 BL vs X Fit Coefficients: Cn sigCn n (Tm/m^n) ----- --------+-------- 0 0.00579 0.00000 Compute Fit Residuals: X0 BL sigBL BLfit BL-BLfit (m) (Tm) (Tm) (Tm) (Tm) ----------- --------+-------- -------- -------- 0.0000000 0.00579 0.00000 0.00579 0.00000 16:01:44 Magnet Strength Calculation Harmonic Number = 1 Number Of X0 Positions = 1 X0 BL sigBL (m) (Tm) (Tm) -------- --------+-------- 0.0000 0.00579 0.00000 SL = 0.0057902 +- 0.0000022 Tm^2/m^n 16:01:44 Magnet Strength Measurement Harmonic Number = 1 SL = 0.0057902 +- 0.0000022 Tm^2/m^n 16:01:47 Position moved: 0.000000 m 16:01:47 Magnet system MAIN current ramping to 30.0000 A... 16:01:59 Magnet Current for MAIN, Imag = 30.0287 A Vtrans = 0.300340 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:02:00 MAIN (V) Voltage Measurement, V = 1.179285 V 16:02:03 Temperature Measurement, Probe Number = 0, T = 25.974000 C 16:02:04 Temperature Measurement, Probe Number = 1, T = 24.098000 C 16:02:05 Temperature Measurement, Probe Number = 2, T = 27.325000 C 16:02:05 Temperature Measurement, Probe Number = 3, T = 28.689000 C 16:02:06 Temperature Measurement, Probe Number = 4, T = 28.692000 C 16:02:06 Temperature Measurement, Probe Number = 5, T = 24.437000 C 16:02:07 Sensor Voltage Measurement, Sensor Number = 0, V = 2.653437 V 16:02:07 Sensor Measurement, Sensor Number = 0, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 2.653437 (V, A, ohm, ...), Measured Value = 124.007745 (C, atm, Hz, ...) 16:02:08 Sensor Voltage Measurement, Sensor Number = 1, V = 1.156023 V 16:02:08 Sensor Measurement, Sensor Number = 1, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 1.156023 (V, A, ohm, ...), Measured Value = 11.701717 (C, atm, Hz, ...) 16:02:08 Magnet Test Parameter Measurements P0 = MAIN, Value = 30.03 P1 = MAIN (V), Value = 1.18 P2 = Yoke, Value = 25.97 P3 = Bottom Coil, Value = 24.10 P4 = Top Coil, Value = 27.32 P5 = Water In, Value = 28.69 P6 = Water Out, Value = 28.69 P7 = Ambient, Value = 24.44 P8 = Supply Pres(psi), Value = 124.01 P9 = Return Pres(psi), Value = 11.70 16:02:08 Magnetic Field Measurement Measuring the magnet strength... 16:02:08 Integrated Voltage Measurement... 16:02:08 Magnet Current for MAIN, Imag = 30.0286 A Vtrans = 0.300339 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:02:17 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.001145514 Vs 16:02:24 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.001160445 Vs 16:02:25 Magnet Current for MAIN, Imag = 30.0286 A Vtrans = 0.300340 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:02:33 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.001143673 Vs 16:02:41 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.001162253 Vs 16:02:41 Magnet Current for MAIN, Imag = 30.0288 A Vtrans = 0.300341 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:02:49 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.001146010 Vs 16:02:56 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.001160945 Vs 16:02:57 Magnet Current for MAIN, Imag = 30.0285 A Vtrans = 0.300338 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:03:05 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.001146541 Vs 16:03:13 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.001161030 Vs 16:03:13 Integrated Voltage Average and RMS of Measurements X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.001153301 +- 0.000000348 Vs 16:03:13 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0000000 m, dx = 0.0100000 m, Nturns = 10.0 VT = 0.001153301 +- 0.000000348 Vs BL = 0.01153301 +- 0.00000348 Tm 16:03:16 Position moved: 0.000000 m 16:03:16 BL vs X Fit Coefficients: Cn sigCn n (Tm/m^n) ----- --------+-------- 0 0.01153 0.00000 Compute Fit Residuals: X0 BL sigBL BLfit BL-BLfit (m) (Tm) (Tm) (Tm) (Tm) ----------- --------+-------- -------- -------- 0.0000000 0.01153 0.00000 0.01153 0.00000 16:03:16 Magnet Strength Calculation Harmonic Number = 1 Number Of X0 Positions = 1 X0 BL sigBL (m) (Tm) (Tm) -------- --------+-------- 0.0000 0.01153 0.00000 SL = 0.0115330 +- 0.0000035 Tm^2/m^n 16:03:16 Magnet Strength Measurement Harmonic Number = 1 SL = 0.0115330 +- 0.0000035 Tm^2/m^n Remeasure 16:03:16 Magnetic Field Measurement Measuring the magnet strength... 16:03:16 Integrated Voltage Measurement... 16:03:17 Magnet Current for MAIN, Imag = 30.0285 A Vtrans = 0.300339 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:03:26 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.001145270 Vs 16:03:33 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.001161714 Vs 16:03:34 Magnet Current for MAIN, Imag = 30.0286 A Vtrans = 0.300339 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:03:42 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.001145524 Vs 16:03:49 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.001161675 Vs 16:03:50 Magnet Current for MAIN, Imag = 30.0286 A Vtrans = 0.300340 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:03:58 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.001145659 Vs 16:04:05 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.001161747 Vs 16:04:06 Magnet Current for MAIN, Imag = 30.0283 A Vtrans = 0.300337 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:04:14 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.001145160 Vs 16:04:21 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.001161524 Vs 16:04:21 Integrated Voltage Average and RMS of Measurements X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.001153534 +- 0.000000134 Vs 16:04:21 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0000000 m, dx = 0.0100000 m, Nturns = 10.0 VT = 0.001153534 +- 0.000000134 Vs BL = 0.01153534 +- 0.00000134 Tm 16:04:25 Position moved: 0.000000 m 16:04:25 BL vs X Fit Coefficients: Cn sigCn n (Tm/m^n) ----- --------+-------- 0 0.01154 0.00000 Compute Fit Residuals: X0 BL sigBL BLfit BL-BLfit (m) (Tm) (Tm) (Tm) (Tm) ----------- --------+-------- -------- -------- 0.0000000 0.01154 0.00000 0.01154 0.00000 16:04:25 Magnet Strength Calculation Harmonic Number = 1 Number Of X0 Positions = 1 X0 BL sigBL (m) (Tm) (Tm) -------- --------+-------- 0.0000 0.01154 0.00000 SL = 0.0115353 +- 0.0000013 Tm^2/m^n 16:04:25 Magnet Strength Measurement Harmonic Number = 1 SL = 0.0115353 +- 0.0000013 Tm^2/m^n 16:04:28 Position moved: 0.000000 m 16:04:28 Magnet system MAIN current ramping to 45.0000 A... 16:04:40 Magnet Current for MAIN, Imag = 44.9811 A Vtrans = 0.449865 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:04:41 MAIN (V) Voltage Measurement, V = 1.766774 V 16:04:44 Temperature Measurement, Probe Number = 0, T = 25.977000 C 16:04:45 Temperature Measurement, Probe Number = 1, T = 24.035000 C 16:04:46 Temperature Measurement, Probe Number = 2, T = 27.376000 C 16:04:46 Temperature Measurement, Probe Number = 3, T = 28.714000 C 16:04:47 Temperature Measurement, Probe Number = 4, T = 28.697000 C 16:04:47 Temperature Measurement, Probe Number = 5, T = 24.409000 C 16:04:48 Sensor Voltage Measurement, Sensor Number = 0, V = 2.641720 V 16:04:48 Sensor Measurement, Sensor Number = 0, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 2.641720 (V, A, ohm, ...), Measured Value = 123.129030 (C, atm, Hz, ...) 16:04:49 Sensor Voltage Measurement, Sensor Number = 1, V = 1.153665 V 16:04:49 Sensor Measurement, Sensor Number = 1, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 1.153665 (V, A, ohm, ...), Measured Value = 11.524905 (C, atm, Hz, ...) 16:04:49 Magnet Test Parameter Measurements P0 = MAIN, Value = 44.98 P1 = MAIN (V), Value = 1.77 P2 = Yoke, Value = 25.98 P3 = Bottom Coil, Value = 24.04 P4 = Top Coil, Value = 27.38 P5 = Water In, Value = 28.71 P6 = Water Out, Value = 28.70 P7 = Ambient, Value = 24.41 P8 = Supply Pres(psi), Value = 123.13 P9 = Return Pres(psi), Value = 11.52 16:04:49 Magnetic Field Measurement Measuring the magnet strength... 16:04:49 Integrated Voltage Measurement... 16:04:49 Magnet Current for MAIN, Imag = 44.9812 A Vtrans = 0.449865 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:04:58 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.001719025 Vs 16:05:06 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.001733623 Vs 16:05:07 Magnet Current for MAIN, Imag = 44.9810 A Vtrans = 0.449864 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:05:14 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.001718449 Vs 16:05:22 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.001732850 Vs 16:05:23 Magnet Current for MAIN, Imag = 44.9811 A Vtrans = 0.449864 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:05:30 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.001719504 Vs 16:05:38 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.001734294 Vs 16:05:39 Magnet Current for MAIN, Imag = 44.9812 A Vtrans = 0.449866 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:05:46 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.001717987 Vs 16:05:54 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.001734601 Vs 16:05:54 Integrated Voltage Average and RMS of Measurements X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.001726292 +- 0.000000442 Vs 16:05:54 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0000000 m, dx = 0.0100000 m, Nturns = 10.0 VT = 0.001726292 +- 0.000000442 Vs BL = 0.01726292 +- 0.00000442 Tm 16:05:58 Position moved: 0.000000 m 16:05:58 BL vs X Fit Coefficients: Cn sigCn n (Tm/m^n) ----- --------+-------- 0 0.01726 0.00000 Compute Fit Residuals: X0 BL sigBL BLfit BL-BLfit (m) (Tm) (Tm) (Tm) (Tm) ----------- --------+-------- -------- -------- 0.0000000 0.01726 0.00000 0.01726 0.00000 16:05:58 Magnet Strength Calculation Harmonic Number = 1 Number Of X0 Positions = 1 X0 BL sigBL (m) (Tm) (Tm) -------- --------+-------- 0.0000 0.01726 0.00000 SL = 0.0172629 +- 0.0000044 Tm^2/m^n 16:05:58 Magnet Strength Measurement Harmonic Number = 1 SL = 0.0172629 +- 0.0000044 Tm^2/m^n Remeasure 16:05:58 Magnetic Field Measurement Measuring the magnet strength... 16:05:58 Integrated Voltage Measurement... 16:05:58 Magnet Current for MAIN, Imag = 44.9810 A Vtrans = 0.449863 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:06:07 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.001720396 Vs 16:06:15 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.001731724 Vs 16:06:15 Magnet Current for MAIN, Imag = 44.9811 A Vtrans = 0.449864 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:06:23 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.001720086 Vs 16:06:31 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.001731607 Vs 16:06:31 Magnet Current for MAIN, Imag = 44.9813 A Vtrans = 0.449866 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:06:39 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.001719216 Vs 16:06:46 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.001734466 Vs 16:06:47 Magnet Current for MAIN, Imag = 44.9813 A Vtrans = 0.449867 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:06:55 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.001718318 Vs 16:07:02 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.001733942 Vs 16:07:02 Integrated Voltage Average and RMS of Measurements X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.001726219 +- 0.000000374 Vs 16:07:02 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0000000 m, dx = 0.0100000 m, Nturns = 10.0 VT = 0.001726219 +- 0.000000374 Vs BL = 0.01726219 +- 0.00000374 Tm 16:07:06 Position moved: 0.000000 m 16:07:06 BL vs X Fit Coefficients: Cn sigCn n (Tm/m^n) ----- --------+-------- 0 0.01726 0.00000 Compute Fit Residuals: X0 BL sigBL BLfit BL-BLfit (m) (Tm) (Tm) (Tm) (Tm) ----------- --------+-------- -------- -------- 0.0000000 0.01726 0.00000 0.01726 0.00000 16:07:06 Magnet Strength Calculation Harmonic Number = 1 Number Of X0 Positions = 1 X0 BL sigBL (m) (Tm) (Tm) -------- --------+-------- 0.0000 0.01726 0.00000 SL = 0.0172622 +- 0.0000037 Tm^2/m^n 16:07:06 Magnet Strength Measurement Harmonic Number = 1 SL = 0.0172622 +- 0.0000037 Tm^2/m^n 16:07:09 Position moved: 0.000000 m 16:07:09 Magnet system MAIN current ramping to 60.0000 A... 16:07:21 Magnet Current for MAIN, Imag = 60.0071 A Vtrans = 0.600124 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:07:22 MAIN (V) Voltage Measurement, V = 2.357592 V 16:07:26 Temperature Measurement, Probe Number = 0, T = 25.969000 C 16:07:26 Temperature Measurement, Probe Number = 1, T = 24.037000 C 16:07:27 Temperature Measurement, Probe Number = 2, T = 27.304000 C 16:07:27 Temperature Measurement, Probe Number = 3, T = 28.770000 C 16:07:28 Temperature Measurement, Probe Number = 4, T = 28.678000 C 16:07:29 Temperature Measurement, Probe Number = 5, T = 24.355000 C 16:07:29 Sensor Voltage Measurement, Sensor Number = 0, V = 2.651467 V 16:07:29 Sensor Measurement, Sensor Number = 0, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 2.651467 (V, A, ohm, ...), Measured Value = 123.860062 (C, atm, Hz, ...) 16:07:30 Sensor Voltage Measurement, Sensor Number = 1, V = 1.155821 V 16:07:30 Sensor Measurement, Sensor Number = 1, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 1.155821 (V, A, ohm, ...), Measured Value = 11.686545 (C, atm, Hz, ...) 16:07:30 Magnet Test Parameter Measurements P0 = MAIN, Value = 60.01 P1 = MAIN (V), Value = 2.36 P2 = Yoke, Value = 25.97 P3 = Bottom Coil, Value = 24.04 P4 = Top Coil, Value = 27.30 P5 = Water In, Value = 28.77 P6 = Water Out, Value = 28.68 P7 = Ambient, Value = 24.36 P8 = Supply Pres(psi), Value = 123.86 P9 = Return Pres(psi), Value = 11.69 16:07:30 Magnetic Field Measurement Measuring the magnet strength... 16:07:30 Integrated Voltage Measurement... 16:07:30 Magnet Current for MAIN, Imag = 60.0066 A Vtrans = 0.600120 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:07:39 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.002293422 Vs 16:07:47 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.002311049 Vs 16:07:48 Magnet Current for MAIN, Imag = 60.0067 A Vtrans = 0.600121 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:07:55 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.002293040 Vs 16:08:03 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.002310285 Vs 16:08:04 Magnet Current for MAIN, Imag = 60.0066 A Vtrans = 0.600119 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:08:11 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.002293668 Vs 16:08:19 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.002311344 Vs 16:08:20 Magnet Current for MAIN, Imag = 60.0066 A Vtrans = 0.600119 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:08:27 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.002294116 Vs 16:08:35 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.002311264 Vs 16:08:35 Integrated Voltage Average and RMS of Measurements X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.002302273 +- 0.000000388 Vs 16:08:35 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0000000 m, dx = 0.0100000 m, Nturns = 10.0 VT = 0.002302273 +- 0.000000388 Vs BL = 0.02302273 +- 0.00000388 Tm 16:08:39 Position moved: 0.000000 m 16:08:39 BL vs X Fit Coefficients: Cn sigCn n (Tm/m^n) ----- --------+-------- 0 0.02302 0.00000 Compute Fit Residuals: X0 BL sigBL BLfit BL-BLfit (m) (Tm) (Tm) (Tm) (Tm) ----------- --------+-------- -------- -------- 0.0000000 0.02302 0.00000 0.02302 0.00000 16:08:39 Magnet Strength Calculation Harmonic Number = 1 Number Of X0 Positions = 1 X0 BL sigBL (m) (Tm) (Tm) -------- --------+-------- 0.0000 0.02302 0.00000 SL = 0.0230227 +- 0.0000039 Tm^2/m^n 16:08:39 Magnet Strength Measurement Harmonic Number = 1 SL = 0.0230227 +- 0.0000039 Tm^2/m^n 16:08:41 Position moved: 0.000000 m 16:08:41 Magnet system MAIN current ramping to 75.0000 A... 16:08:54 Magnet Current for MAIN, Imag = 75.0360 A Vtrans = 0.750414 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:08:54 MAIN (V) Voltage Measurement, V = 2.949092 V 16:08:58 Temperature Measurement, Probe Number = 0, T = 25.963000 C 16:08:58 Temperature Measurement, Probe Number = 1, T = 24.050000 C 16:08:59 Temperature Measurement, Probe Number = 2, T = 27.292000 C 16:09:00 Temperature Measurement, Probe Number = 3, T = 28.850000 C 16:09:00 Temperature Measurement, Probe Number = 4, T = 28.689000 C 16:09:01 Temperature Measurement, Probe Number = 5, T = 24.396000 C 16:09:01 Sensor Voltage Measurement, Sensor Number = 0, V = 2.655372 V 16:09:01 Sensor Measurement, Sensor Number = 0, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 2.655372 (V, A, ohm, ...), Measured Value = 124.152908 (C, atm, Hz, ...) 16:09:02 Sensor Voltage Measurement, Sensor Number = 1, V = 1.156223 V 16:09:02 Sensor Measurement, Sensor Number = 1, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 1.156223 (V, A, ohm, ...), Measured Value = 11.716703 (C, atm, Hz, ...) 16:09:02 Magnet Test Parameter Measurements P0 = MAIN, Value = 75.04 P1 = MAIN (V), Value = 2.95 P2 = Yoke, Value = 25.96 P3 = Bottom Coil, Value = 24.05 P4 = Top Coil, Value = 27.29 P5 = Water In, Value = 28.85 P6 = Water Out, Value = 28.69 P7 = Ambient, Value = 24.40 P8 = Supply Pres(psi), Value = 124.15 P9 = Return Pres(psi), Value = 11.72 16:09:02 Magnetic Field Measurement Measuring the magnet strength... 16:09:02 Integrated Voltage Measurement... 16:09:03 Magnet Current for MAIN, Imag = 75.0362 A Vtrans = 0.750416 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:09:11 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.002870801 Vs 16:09:19 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.002885391 Vs 16:09:20 Magnet Current for MAIN, Imag = 75.0365 A Vtrans = 0.750419 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:09:27 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.002870000 Vs 16:09:35 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.002885554 Vs 16:09:36 Magnet Current for MAIN, Imag = 75.0364 A Vtrans = 0.750417 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:09:43 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.002868965 Vs 16:09:51 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.002887544 Vs 16:09:51 Magnet Current for MAIN, Imag = 75.0363 A Vtrans = 0.750417 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:09:59 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.002871139 Vs 16:10:07 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.002885956 Vs 16:10:07 Integrated Voltage Average and RMS of Measurements X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.002878169 +- 0.000000278 Vs 16:10:07 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0000000 m, dx = 0.0100000 m, Nturns = 10.0 VT = 0.002878169 +- 0.000000278 Vs BL = 0.02878169 +- 0.00000278 Tm 16:10:11 Position moved: 0.000000 m 16:10:11 BL vs X Fit Coefficients: Cn sigCn n (Tm/m^n) ----- --------+-------- 0 0.02878 0.00000 Compute Fit Residuals: X0 BL sigBL BLfit BL-BLfit (m) (Tm) (Tm) (Tm) (Tm) ----------- --------+-------- -------- -------- 0.0000000 0.02878 0.00000 0.02878 0.00000 16:10:11 Magnet Strength Calculation Harmonic Number = 1 Number Of X0 Positions = 1 X0 BL sigBL (m) (Tm) (Tm) -------- --------+-------- 0.0000 0.02878 0.00000 SL = 0.0287817 +- 0.0000028 Tm^2/m^n 16:10:11 Magnet Strength Measurement Harmonic Number = 1 SL = 0.0287817 +- 0.0000028 Tm^2/m^n 16:10:13 Position moved: 0.000000 m 16:10:13 Magnet system MAIN current ramping to 90.0000 A... 16:10:25 Magnet Current for MAIN, Imag = 89.9906 A Vtrans = 0.899959 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:10:26 MAIN (V) Voltage Measurement, V = 3.538194 V 16:10:30 Temperature Measurement, Probe Number = 0, T = 25.968000 C 16:10:30 Temperature Measurement, Probe Number = 1, T = 24.012000 C 16:10:31 Temperature Measurement, Probe Number = 2, T = 27.369000 C 16:10:31 Temperature Measurement, Probe Number = 3, T = 28.971000 C 16:10:32 Temperature Measurement, Probe Number = 4, T = 28.686000 C 16:10:32 Temperature Measurement, Probe Number = 5, T = 24.360000 C 16:10:33 Sensor Voltage Measurement, Sensor Number = 0, V = 2.652719 V 16:10:33 Sensor Measurement, Sensor Number = 0, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 2.652719 (V, A, ohm, ...), Measured Value = 123.953902 (C, atm, Hz, ...) 16:10:34 Sensor Voltage Measurement, Sensor Number = 1, V = 1.153595 V 16:10:34 Sensor Measurement, Sensor Number = 1, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 1.153595 (V, A, ohm, ...), Measured Value = 11.519662 (C, atm, Hz, ...) 16:10:34 Magnet Test Parameter Measurements P0 = MAIN, Value = 89.99 P1 = MAIN (V), Value = 3.54 P2 = Yoke, Value = 25.97 P3 = Bottom Coil, Value = 24.01 P4 = Top Coil, Value = 27.37 P5 = Water In, Value = 28.97 P6 = Water Out, Value = 28.69 P7 = Ambient, Value = 24.36 P8 = Supply Pres(psi), Value = 123.95 P9 = Return Pres(psi), Value = 11.52 16:10:34 Magnetic Field Measurement Measuring the magnet strength... 16:10:34 Integrated Voltage Measurement... 16:10:34 Magnet Current for MAIN, Imag = 89.9906 A Vtrans = 0.899959 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:10:43 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.003442681 Vs 16:10:51 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.003461530 Vs 16:10:51 Magnet Current for MAIN, Imag = 89.9907 A Vtrans = 0.899960 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:10:59 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.003440064 Vs 16:11:06 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.003461745 Vs 16:11:07 Magnet Current for MAIN, Imag = 89.9903 A Vtrans = 0.899957 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:11:15 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.003441896 Vs 16:11:22 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.003460969 Vs 16:11:23 Magnet Current for MAIN, Imag = 89.9905 A Vtrans = 0.899958 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:11:31 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.003442317 Vs 16:11:39 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.003458203 Vs 16:11:39 Integrated Voltage Average and RMS of Measurements X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.003451176 +- 0.000000679 Vs 16:11:39 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0000000 m, dx = 0.0100000 m, Nturns = 10.0 VT = 0.003451176 +- 0.000000679 Vs BL = 0.03451176 +- 0.00000679 Tm 16:11:42 Position moved: 0.000000 m 16:11:42 BL vs X Fit Coefficients: Cn sigCn n (Tm/m^n) ----- --------+-------- 0 0.03451 0.00001 Compute Fit Residuals: X0 BL sigBL BLfit BL-BLfit (m) (Tm) (Tm) (Tm) (Tm) ----------- --------+-------- -------- -------- 0.0000000 0.03451 0.00001 0.03451 0.00000 16:11:42 Magnet Strength Calculation Harmonic Number = 1 Number Of X0 Positions = 1 X0 BL sigBL (m) (Tm) (Tm) -------- --------+-------- 0.0000 0.03451 0.00001 SL = 0.0345118 +- 0.0000068 Tm^2/m^n 16:11:42 Magnet Strength Measurement Harmonic Number = 1 SL = 0.0345118 +- 0.0000068 Tm^2/m^n 16:11:45 Position moved: 0.000000 m 16:11:45 Magnet system MAIN current ramping to 105.0000 A... 16:11:57 Magnet Current for MAIN, Imag = 105.0181 A Vtrans = 1.050235 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:11:58 MAIN (V) Voltage Measurement, V = 4.131342 V 16:12:01 Temperature Measurement, Probe Number = 0, T = 25.956000 C 16:12:02 Temperature Measurement, Probe Number = 1, T = 24.017000 C 16:12:03 Temperature Measurement, Probe Number = 2, T = 27.455000 C 16:12:03 Temperature Measurement, Probe Number = 3, T = 29.088000 C 16:12:04 Temperature Measurement, Probe Number = 4, T = 28.677000 C 16:12:04 Temperature Measurement, Probe Number = 5, T = 24.384000 C 16:12:05 Sensor Voltage Measurement, Sensor Number = 0, V = 2.662659 V 16:12:05 Sensor Measurement, Sensor Number = 0, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 2.662659 (V, A, ohm, ...), Measured Value = 124.699448 (C, atm, Hz, ...) 16:12:05 Sensor Voltage Measurement, Sensor Number = 1, V = 1.154264 V 16:12:06 Sensor Measurement, Sensor Number = 1, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 1.154264 (V, A, ohm, ...), Measured Value = 11.569763 (C, atm, Hz, ...) 16:12:06 Magnet Test Parameter Measurements P0 = MAIN, Value = 105.02 P1 = MAIN (V), Value = 4.13 P2 = Yoke, Value = 25.96 P3 = Bottom Coil, Value = 24.02 P4 = Top Coil, Value = 27.45 P5 = Water In, Value = 29.09 P6 = Water Out, Value = 28.68 P7 = Ambient, Value = 24.38 P8 = Supply Pres(psi), Value = 124.70 P9 = Return Pres(psi), Value = 11.57 16:12:06 Magnetic Field Measurement Measuring the magnet strength... 16:12:06 Integrated Voltage Measurement... 16:12:06 Magnet Current for MAIN, Imag = 105.0182 A Vtrans = 1.050236 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:12:15 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.004017858 Vs 16:12:23 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.004035755 Vs 16:12:23 Magnet Current for MAIN, Imag = 105.0183 A Vtrans = 1.050237 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:12:31 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.004021707 Vs 16:12:39 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.004035328 Vs 16:12:39 Magnet Current for MAIN, Imag = 105.0183 A Vtrans = 1.050237 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:12:47 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.004017333 Vs 16:12:55 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.004034432 Vs 16:12:55 Magnet Current for MAIN, Imag = 105.0181 A Vtrans = 1.050234 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:13:03 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.004018018 Vs 16:13:10 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.004033261 Vs 16:13:10 Integrated Voltage Average and RMS of Measurements X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.004026711 +- 0.000001130 Vs 16:13:10 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0000000 m, dx = 0.0100000 m, Nturns = 10.0 VT = 0.004026711 +- 0.000001130 Vs BL = 0.04026711 +- 0.00001130 Tm 16:13:14 Position moved: 0.000000 m 16:13:14 BL vs X Fit Coefficients: Cn sigCn n (Tm/m^n) ----- --------+-------- 0 0.04027 0.00001 Compute Fit Residuals: X0 BL sigBL BLfit BL-BLfit (m) (Tm) (Tm) (Tm) (Tm) ----------- --------+-------- -------- -------- 0.0000000 0.04027 0.00001 0.04027 0.00000 16:13:14 Magnet Strength Calculation Harmonic Number = 1 Number Of X0 Positions = 1 X0 BL sigBL (m) (Tm) (Tm) -------- --------+-------- 0.0000 0.04027 0.00001 SL = 0.0402671 +- 0.0000113 Tm^2/m^n 16:13:14 Magnet Strength Measurement Harmonic Number = 1 SL = 0.0402671 +- 0.0000113 Tm^2/m^n Remeasure 16:13:14 Magnetic Field Measurement Measuring the magnet strength... 16:13:14 Integrated Voltage Measurement... 16:13:15 Magnet Current for MAIN, Imag = 105.0184 A Vtrans = 1.050237 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:13:24 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.004020773 Vs 16:13:31 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.004033943 Vs 16:13:32 Magnet Current for MAIN, Imag = 105.0183 A Vtrans = 1.050236 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:13:40 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.004019138 Vs 16:13:47 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.004031728 Vs 16:13:48 Magnet Current for MAIN, Imag = 105.0180 A Vtrans = 1.050233 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:13:56 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.004018902 Vs 16:14:03 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.004034333 Vs 16:14:04 Magnet Current for MAIN, Imag = 105.0180 A Vtrans = 1.050234 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:14:12 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.004020513 Vs 16:14:19 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.004035968 Vs 16:14:19 Integrated Voltage Average and RMS of Measurements X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.004026912 +- 0.000001029 Vs 16:14:19 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0000000 m, dx = 0.0100000 m, Nturns = 10.0 VT = 0.004026912 +- 0.000001029 Vs BL = 0.04026912 +- 0.00001029 Tm 16:14:23 Position moved: 0.000000 m 16:14:23 BL vs X Fit Coefficients: Cn sigCn n (Tm/m^n) ----- --------+-------- 0 0.04027 0.00001 Compute Fit Residuals: X0 BL sigBL BLfit BL-BLfit (m) (Tm) (Tm) (Tm) (Tm) ----------- --------+-------- -------- -------- 0.0000000 0.04027 0.00001 0.04027 0.00000 16:14:23 Magnet Strength Calculation Harmonic Number = 1 Number Of X0 Positions = 1 X0 BL sigBL (m) (Tm) (Tm) -------- --------+-------- 0.0000 0.04027 0.00001 SL = 0.0402691 +- 0.0000103 Tm^2/m^n 16:14:23 Magnet Strength Measurement Harmonic Number = 1 SL = 0.0402691 +- 0.0000103 Tm^2/m^n 16:14:26 Position moved: 0.000000 m 16:14:26 Magnet system MAIN current ramping to 120.0000 A... 16:14:38 Magnet Current for MAIN, Imag = 120.0370 A Vtrans = 1.200423 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:14:39 MAIN (V) Voltage Measurement, V = 4.725224 V 16:14:42 Temperature Measurement, Probe Number = 0, T = 25.958000 C 16:14:43 Temperature Measurement, Probe Number = 1, T = 24.029000 C 16:14:43 Temperature Measurement, Probe Number = 2, T = 27.592000 C 16:14:44 Temperature Measurement, Probe Number = 3, T = 29.320000 C 16:14:45 Temperature Measurement, Probe Number = 4, T = 28.641000 C 16:14:45 Temperature Measurement, Probe Number = 5, T = 24.368000 C 16:14:46 Sensor Voltage Measurement, Sensor Number = 0, V = 2.650533 V 16:14:46 Sensor Measurement, Sensor Number = 0, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 2.650533 (V, A, ohm, ...), Measured Value = 123.790012 (C, atm, Hz, ...) 16:14:46 Sensor Voltage Measurement, Sensor Number = 1, V = 1.155471 V 16:14:46 Sensor Measurement, Sensor Number = 1, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 1.155471 (V, A, ohm, ...), Measured Value = 11.660325 (C, atm, Hz, ...) 16:14:46 Magnet Test Parameter Measurements P0 = MAIN, Value = 120.04 P1 = MAIN (V), Value = 4.73 P2 = Yoke, Value = 25.96 P3 = Bottom Coil, Value = 24.03 P4 = Top Coil, Value = 27.59 P5 = Water In, Value = 29.32 P6 = Water Out, Value = 28.64 P7 = Ambient, Value = 24.37 P8 = Supply Pres(psi), Value = 123.79 P9 = Return Pres(psi), Value = 11.66 16:14:46 Magnetic Field Measurement Measuring the magnet strength... 16:14:46 Integrated Voltage Measurement... 16:14:47 Magnet Current for MAIN, Imag = 120.0365 A Vtrans = 1.200418 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:14:56 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.004592692 Vs 16:15:03 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.004607878 Vs 16:15:04 Magnet Current for MAIN, Imag = 120.0370 A Vtrans = 1.200423 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:15:12 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.004593231 Vs 16:15:19 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.004607314 Vs 16:15:20 Magnet Current for MAIN, Imag = 120.0370 A Vtrans = 1.200423 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:15:28 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.004592938 Vs 16:15:35 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.004610044 Vs 16:15:36 Magnet Current for MAIN, Imag = 120.0371 A Vtrans = 1.200424 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:15:44 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.004593466 Vs 16:15:51 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.004606849 Vs 16:15:51 Integrated Voltage Average and RMS of Measurements X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.004600552 +- 0.000000545 Vs 16:15:51 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0000000 m, dx = 0.0100000 m, Nturns = 10.0 VT = 0.004600552 +- 0.000000545 Vs BL = 0.04600552 +- 0.00000545 Tm 16:15:55 Position moved: 0.000000 m 16:15:55 BL vs X Fit Coefficients: Cn sigCn n (Tm/m^n) ----- --------+-------- 0 0.04601 0.00001 Compute Fit Residuals: X0 BL sigBL BLfit BL-BLfit (m) (Tm) (Tm) (Tm) (Tm) ----------- --------+-------- -------- -------- 0.0000000 0.04601 0.00001 0.04601 0.00000 16:15:55 Magnet Strength Calculation Harmonic Number = 1 Number Of X0 Positions = 1 X0 BL sigBL (m) (Tm) (Tm) -------- --------+-------- 0.0000 0.04601 0.00001 SL = 0.0460055 +- 0.0000054 Tm^2/m^n 16:15:55 Magnet Strength Measurement Harmonic Number = 1 SL = 0.0460055 +- 0.0000054 Tm^2/m^n 16:15:58 Position moved: 0.000000 m 16:15:58 Magnet system MAIN current ramping to 135.0000 A... 16:16:10 Magnet Current for MAIN, Imag = 134.9910 A Vtrans = 1.349964 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:16:11 MAIN (V) Voltage Measurement, V = 5.317768 V 16:16:14 Temperature Measurement, Probe Number = 0, T = 25.964000 C 16:16:15 Temperature Measurement, Probe Number = 1, T = 24.048000 C 16:16:15 Temperature Measurement, Probe Number = 2, T = 27.674000 C 16:16:16 Temperature Measurement, Probe Number = 3, T = 29.525000 C 16:16:17 Temperature Measurement, Probe Number = 4, T = 28.670000 C 16:16:17 Temperature Measurement, Probe Number = 5, T = 24.385000 C 16:16:18 Sensor Voltage Measurement, Sensor Number = 0, V = 2.651215 V 16:16:18 Sensor Measurement, Sensor Number = 0, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 2.651215 (V, A, ohm, ...), Measured Value = 123.841148 (C, atm, Hz, ...) 16:16:18 Sensor Voltage Measurement, Sensor Number = 1, V = 1.156899 V 16:16:18 Sensor Measurement, Sensor Number = 1, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 1.156899 (V, A, ohm, ...), Measured Value = 11.767462 (C, atm, Hz, ...) 16:16:18 Magnet Test Parameter Measurements P0 = MAIN, Value = 134.99 P1 = MAIN (V), Value = 5.32 P2 = Yoke, Value = 25.96 P3 = Bottom Coil, Value = 24.05 P4 = Top Coil, Value = 27.67 P5 = Water In, Value = 29.52 P6 = Water Out, Value = 28.67 P7 = Ambient, Value = 24.39 P8 = Supply Pres(psi), Value = 123.84 P9 = Return Pres(psi), Value = 11.77 16:16:18 Magnetic Field Measurement Measuring the magnet strength... 16:16:18 Integrated Voltage Measurement... 16:16:19 Magnet Current for MAIN, Imag = 134.9905 A Vtrans = 1.349959 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:16:28 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.005163363 Vs 16:16:35 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.005181941 Vs 16:16:36 Magnet Current for MAIN, Imag = 134.9909 A Vtrans = 1.349962 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:16:44 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.005166340 Vs 16:16:51 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.005181651 Vs 16:16:52 Magnet Current for MAIN, Imag = 134.9913 A Vtrans = 1.349966 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:17:00 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.005161007 Vs 16:17:07 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.005181148 Vs 16:17:08 Magnet Current for MAIN, Imag = 134.9909 A Vtrans = 1.349962 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:17:16 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.005162451 Vs 16:17:24 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.005184163 Vs 16:17:24 Integrated Voltage Average and RMS of Measurements X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.005172758 +- 0.000001081 Vs 16:17:24 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0000000 m, dx = 0.0100000 m, Nturns = 10.0 VT = 0.005172758 +- 0.000001081 Vs BL = 0.05172758 +- 0.00001081 Tm 16:17:27 Position moved: 0.000000 m 16:17:27 BL vs X Fit Coefficients: Cn sigCn n (Tm/m^n) ----- --------+-------- 0 0.05173 0.00001 Compute Fit Residuals: X0 BL sigBL BLfit BL-BLfit (m) (Tm) (Tm) (Tm) (Tm) ----------- --------+-------- -------- -------- 0.0000000 0.05173 0.00001 0.05173 0.00000 16:17:28 Magnet Strength Calculation Harmonic Number = 1 Number Of X0 Positions = 1 X0 BL sigBL (m) (Tm) (Tm) -------- --------+-------- 0.0000 0.05173 0.00001 SL = 0.0517276 +- 0.0000108 Tm^2/m^n 16:17:28 Magnet Strength Measurement Harmonic Number = 1 SL = 0.0517276 +- 0.0000108 Tm^2/m^n 16:17:30 Position moved: 0.000000 m 16:17:30 Magnet system MAIN current ramping to 150.0000 A... 16:17:42 Magnet Current for MAIN, Imag = 150.0259 A Vtrans = 1.500312 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:17:43 MAIN (V) Voltage Measurement, V = 5.914828 V 16:17:47 Temperature Measurement, Probe Number = 0, T = 25.968000 C 16:17:47 Temperature Measurement, Probe Number = 1, T = 24.010000 C 16:17:48 Temperature Measurement, Probe Number = 2, T = 27.726000 C 16:17:48 Temperature Measurement, Probe Number = 3, T = 29.790000 C 16:17:49 Temperature Measurement, Probe Number = 4, T = 28.676000 C 16:17:50 Temperature Measurement, Probe Number = 5, T = 24.374000 C 16:17:50 Sensor Voltage Measurement, Sensor Number = 0, V = 2.641866 V 16:17:50 Sensor Measurement, Sensor Number = 0, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 2.641866 (V, A, ohm, ...), Measured Value = 123.139987 (C, atm, Hz, ...) 16:17:51 Sensor Voltage Measurement, Sensor Number = 1, V = 1.153863 V 16:17:51 Sensor Measurement, Sensor Number = 1, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 1.153863 (V, A, ohm, ...), Measured Value = 11.539702 (C, atm, Hz, ...) 16:17:51 Magnet Test Parameter Measurements P0 = MAIN, Value = 150.03 P1 = MAIN (V), Value = 5.91 P2 = Yoke, Value = 25.97 P3 = Bottom Coil, Value = 24.01 P4 = Top Coil, Value = 27.73 P5 = Water In, Value = 29.79 P6 = Water Out, Value = 28.68 P7 = Ambient, Value = 24.37 P8 = Supply Pres(psi), Value = 123.14 P9 = Return Pres(psi), Value = 11.54 16:17:51 Magnetic Field Measurement Measuring the magnet strength... 16:17:51 Integrated Voltage Measurement... 16:17:52 Magnet Current for MAIN, Imag = 150.0262 A Vtrans = 1.500316 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:18:00 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.005736552 Vs 16:18:08 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.005753829 Vs 16:18:09 Magnet Current for MAIN, Imag = 150.0264 A Vtrans = 1.500317 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:18:16 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.005736386 Vs 16:18:24 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.005755243 Vs 16:18:25 Magnet Current for MAIN, Imag = 150.0269 A Vtrans = 1.500322 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:18:32 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.005735732 Vs 16:18:40 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.005755349 Vs 16:18:41 Magnet Current for MAIN, Imag = 150.0265 A Vtrans = 1.500318 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:18:48 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.005736402 Vs 16:18:56 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.005753306 Vs 16:18:56 Integrated Voltage Average and RMS of Measurements X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.005745350 +- 0.000000362 Vs 16:18:56 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0000000 m, dx = 0.0100000 m, Nturns = 10.0 VT = 0.005745350 +- 0.000000362 Vs BL = 0.05745350 +- 0.00000362 Tm 16:19:00 Position moved: 0.000000 m 16:19:00 BL vs X Fit Coefficients: Cn sigCn n (Tm/m^n) ----- --------+-------- 0 0.05745 0.00000 Compute Fit Residuals: X0 BL sigBL BLfit BL-BLfit (m) (Tm) (Tm) (Tm) (Tm) ----------- --------+-------- -------- -------- 0.0000000 0.05745 0.00000 0.05745 0.00000 16:19:00 Magnet Strength Calculation Harmonic Number = 1 Number Of X0 Positions = 1 X0 BL sigBL (m) (Tm) (Tm) -------- --------+-------- 0.0000 0.05745 0.00000 SL = 0.0574535 +- 0.0000036 Tm^2/m^n 16:19:00 Magnet Strength Measurement Harmonic Number = 1 SL = 0.0574535 +- 0.0000036 Tm^2/m^n 16:19:03 Position moved: 0.000000 m 16:19:03 Magnet system MAIN current ramping to 165.0000 A... 16:19:15 Magnet Current for MAIN, Imag = 164.9768 A Vtrans = 1.649821 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:19:15 MAIN (V) Voltage Measurement, V = 6.510154 V 16:19:19 Temperature Measurement, Probe Number = 0, T = 25.980000 C 16:19:19 Temperature Measurement, Probe Number = 1, T = 24.047000 C 16:19:20 Temperature Measurement, Probe Number = 2, T = 27.914000 C 16:19:21 Temperature Measurement, Probe Number = 3, T = 30.092000 C 16:19:21 Temperature Measurement, Probe Number = 4, T = 28.681000 C 16:19:22 Temperature Measurement, Probe Number = 5, T = 24.388000 C 16:19:22 Sensor Voltage Measurement, Sensor Number = 0, V = 2.655522 V 16:19:22 Sensor Measurement, Sensor Number = 0, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 2.655522 (V, A, ohm, ...), Measured Value = 124.164143 (C, atm, Hz, ...) 16:19:23 Sensor Voltage Measurement, Sensor Number = 1, V = 1.156097 V 16:19:23 Sensor Measurement, Sensor Number = 1, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 1.156097 (V, A, ohm, ...), Measured Value = 11.707245 (C, atm, Hz, ...) 16:19:23 Magnet Test Parameter Measurements P0 = MAIN, Value = 164.98 P1 = MAIN (V), Value = 6.51 P2 = Yoke, Value = 25.98 P3 = Bottom Coil, Value = 24.05 P4 = Top Coil, Value = 27.91 P5 = Water In, Value = 30.09 P6 = Water Out, Value = 28.68 P7 = Ambient, Value = 24.39 P8 = Supply Pres(psi), Value = 124.16 P9 = Return Pres(psi), Value = 11.71 16:19:23 Magnetic Field Measurement Measuring the magnet strength... 16:19:23 Integrated Voltage Measurement... 16:19:24 Magnet Current for MAIN, Imag = 164.9769 A Vtrans = 1.649823 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:19:32 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.006306205 Vs 16:19:40 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.006320909 Vs 16:19:41 Magnet Current for MAIN, Imag = 164.9770 A Vtrans = 1.649824 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:19:48 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.006304002 Vs 16:19:56 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.006325358 Vs 16:19:57 Magnet Current for MAIN, Imag = 164.9769 A Vtrans = 1.649823 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:20:04 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.006303808 Vs 16:20:12 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.006324779 Vs 16:20:13 Magnet Current for MAIN, Imag = 164.9773 A Vtrans = 1.649826 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:20:20 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.006304272 Vs 16:20:28 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.006322828 Vs 16:20:28 Integrated Voltage Average and RMS of Measurements X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.006314020 +- 0.000000486 Vs 16:20:28 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0000000 m, dx = 0.0100000 m, Nturns = 10.0 VT = 0.006314020 +- 0.000000486 Vs BL = 0.06314020 +- 0.00000486 Tm 16:20:32 Position moved: 0.000000 m 16:20:32 BL vs X Fit Coefficients: Cn sigCn n (Tm/m^n) ----- --------+-------- 0 0.06314 0.00000 Compute Fit Residuals: X0 BL sigBL BLfit BL-BLfit (m) (Tm) (Tm) (Tm) (Tm) ----------- --------+-------- -------- -------- 0.0000000 0.06314 0.00000 0.06314 0.00000 16:20:32 Magnet Strength Calculation Harmonic Number = 1 Number Of X0 Positions = 1 X0 BL sigBL (m) (Tm) (Tm) -------- --------+-------- 0.0000 0.06314 0.00000 SL = 0.0631402 +- 0.0000049 Tm^2/m^n 16:20:32 Magnet Strength Measurement Harmonic Number = 1 SL = 0.0631402 +- 0.0000049 Tm^2/m^n 16:20:35 Position moved: 0.000000 m 16:20:35 Magnet system MAIN current ramping to 180.0000 A... 16:20:47 Magnet Current for MAIN, Imag = 180.0128 A Vtrans = 1.800181 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:20:47 MAIN (V) Voltage Measurement, V = 7.110351 V 16:20:51 Temperature Measurement, Probe Number = 0, T = 25.988000 C 16:20:52 Temperature Measurement, Probe Number = 1, T = 24.101000 C 16:20:52 Temperature Measurement, Probe Number = 2, T = 28.067000 C 16:20:53 Temperature Measurement, Probe Number = 3, T = 30.410000 C 16:20:53 Temperature Measurement, Probe Number = 4, T = 28.691000 C 16:20:54 Temperature Measurement, Probe Number = 5, T = 24.402000 C 16:20:54 Sensor Voltage Measurement, Sensor Number = 0, V = 2.657036 V 16:20:54 Sensor Measurement, Sensor Number = 0, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 2.657036 (V, A, ohm, ...), Measured Value = 124.277737 (C, atm, Hz, ...) 16:20:55 Sensor Voltage Measurement, Sensor Number = 1, V = 1.156833 V 16:20:55 Sensor Measurement, Sensor Number = 1, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 1.156833 (V, A, ohm, ...), Measured Value = 11.762497 (C, atm, Hz, ...) 16:20:55 Magnet Test Parameter Measurements P0 = MAIN, Value = 180.01 P1 = MAIN (V), Value = 7.11 P2 = Yoke, Value = 25.99 P3 = Bottom Coil, Value = 24.10 P4 = Top Coil, Value = 28.07 P5 = Water In, Value = 30.41 P6 = Water Out, Value = 28.69 P7 = Ambient, Value = 24.40 P8 = Supply Pres(psi), Value = 124.28 P9 = Return Pres(psi), Value = 11.76 16:20:55 Magnetic Field Measurement Measuring the magnet strength... 16:20:55 Integrated Voltage Measurement... 16:20:56 Magnet Current for MAIN, Imag = 180.0132 A Vtrans = 1.800185 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:21:04 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.006874264 Vs 16:21:12 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.006889360 Vs 16:21:13 Magnet Current for MAIN, Imag = 180.0130 A Vtrans = 1.800184 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:21:20 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.006872863 Vs 16:21:28 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.006892318 Vs 16:21:29 Magnet Current for MAIN, Imag = 180.0133 A Vtrans = 1.800186 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:21:36 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.006875052 Vs 16:21:44 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.006888809 Vs 16:21:45 Magnet Current for MAIN, Imag = 180.0129 A Vtrans = 1.800183 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:21:52 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.006876290 Vs 16:22:00 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.006888980 Vs 16:22:00 Integrated Voltage Average and RMS of Measurements X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.006882242 +- 0.000000373 Vs 16:22:00 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0000000 m, dx = 0.0100000 m, Nturns = 10.0 VT = 0.006882242 +- 0.000000373 Vs BL = 0.06882242 +- 0.00000373 Tm 16:22:03 Position moved: 0.000000 m 16:22:03 BL vs X Fit Coefficients: Cn sigCn n (Tm/m^n) ----- --------+-------- 0 0.06882 0.00000 Compute Fit Residuals: X0 BL sigBL BLfit BL-BLfit (m) (Tm) (Tm) (Tm) (Tm) ----------- --------+-------- -------- -------- 0.0000000 0.06882 0.00000 0.06882 0.00000 16:22:03 Magnet Strength Calculation Harmonic Number = 1 Number Of X0 Positions = 1 X0 BL sigBL (m) (Tm) (Tm) -------- --------+-------- 0.0000 0.06882 0.00000 SL = 0.0688224 +- 0.0000037 Tm^2/m^n 16:22:03 Magnet Strength Measurement Harmonic Number = 1 SL = 0.0688224 +- 0.0000037 Tm^2/m^n 16:22:06 Position moved: 0.000000 m 16:22:06 Magnet system MAIN current ramping to 185.0000 A... 16:22:18 Magnet Current for MAIN, Imag = 185.0454 A Vtrans = 1.850507 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:22:18 MAIN (V) Voltage Measurement, V = 7.312169 V 16:22:22 Temperature Measurement, Probe Number = 0, T = 26.005000 C 16:22:22 Temperature Measurement, Probe Number = 1, T = 24.122000 C 16:22:23 Temperature Measurement, Probe Number = 2, T = 28.428000 C 16:22:24 Temperature Measurement, Probe Number = 3, T = 30.762000 C 16:22:24 Temperature Measurement, Probe Number = 4, T = 28.691000 C 16:22:25 Temperature Measurement, Probe Number = 5, T = 24.446000 C 16:22:25 Sensor Voltage Measurement, Sensor Number = 0, V = 2.643927 V 16:22:25 Sensor Measurement, Sensor Number = 0, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 2.643927 (V, A, ohm, ...), Measured Value = 123.294510 (C, atm, Hz, ...) 16:22:26 Sensor Voltage Measurement, Sensor Number = 1, V = 1.156320 V 16:22:26 Sensor Measurement, Sensor Number = 1, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 1.156320 (V, A, ohm, ...), Measured Value = 11.724008 (C, atm, Hz, ...) 16:22:26 Magnet Test Parameter Measurements P0 = MAIN, Value = 185.05 P1 = MAIN (V), Value = 7.31 P2 = Yoke, Value = 26.00 P3 = Bottom Coil, Value = 24.12 P4 = Top Coil, Value = 28.43 P5 = Water In, Value = 30.76 P6 = Water Out, Value = 28.69 P7 = Ambient, Value = 24.45 P8 = Supply Pres(psi), Value = 123.29 P9 = Return Pres(psi), Value = 11.72 16:22:26 Magnetic Field Measurement Measuring the magnet strength... 16:22:26 Integrated Voltage Measurement... 16:22:27 Magnet Current for MAIN, Imag = 185.0455 A Vtrans = 1.850509 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:22:35 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.007062192 Vs 16:22:43 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.007080871 Vs 16:22:44 Magnet Current for MAIN, Imag = 185.0455 A Vtrans = 1.850509 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:22:52 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.007062197 Vs 16:22:59 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.007077460 Vs 16:23:00 Magnet Current for MAIN, Imag = 185.0454 A Vtrans = 1.850507 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:23:08 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.007065224 Vs 16:23:15 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.007076736 Vs 16:23:16 Magnet Current for MAIN, Imag = 185.0455 A Vtrans = 1.850509 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:23:24 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.007062050 Vs 16:23:31 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.007078538 Vs 16:23:31 Integrated Voltage Average and RMS of Measurements X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.007070658 +- 0.000000650 Vs 16:23:31 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0000000 m, dx = 0.0100000 m, Nturns = 10.0 VT = 0.007070658 +- 0.000000650 Vs BL = 0.07070658 +- 0.00000650 Tm 16:23:35 Position moved: 0.000000 m 16:23:35 BL vs X Fit Coefficients: Cn sigCn n (Tm/m^n) ----- --------+-------- 0 0.07071 0.00001 Compute Fit Residuals: X0 BL sigBL BLfit BL-BLfit (m) (Tm) (Tm) (Tm) (Tm) ----------- --------+-------- -------- -------- 0.0000000 0.07071 0.00001 0.07071 0.00000 16:23:35 Magnet Strength Calculation Harmonic Number = 1 Number Of X0 Positions = 1 X0 BL sigBL (m) (Tm) (Tm) -------- --------+-------- 0.0000 0.07071 0.00001 SL = 0.0707066 +- 0.0000065 Tm^2/m^n 16:23:35 Magnet Strength Measurement Harmonic Number = 1 SL = 0.0707066 +- 0.0000065 Tm^2/m^n 16:23:38 Position moved: 0.000000 m 16:23:38 Magnet system MAIN current ramping to 180.0000 A... 16:23:50 Magnet Current for MAIN, Imag = 180.0127 A Vtrans = 1.800180 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:23:50 MAIN (V) Voltage Measurement, V = 7.111313 V 16:23:54 Temperature Measurement, Probe Number = 0, T = 26.011000 C 16:23:54 Temperature Measurement, Probe Number = 1, T = 24.129000 C 16:23:55 Temperature Measurement, Probe Number = 2, T = 28.608000 C 16:23:56 Temperature Measurement, Probe Number = 3, T = 30.960000 C 16:23:56 Temperature Measurement, Probe Number = 4, T = 28.662000 C 16:23:57 Temperature Measurement, Probe Number = 5, T = 24.437000 C 16:23:57 Sensor Voltage Measurement, Sensor Number = 0, V = 2.650808 V 16:23:57 Sensor Measurement, Sensor Number = 0, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 2.650808 (V, A, ohm, ...), Measured Value = 123.810615 (C, atm, Hz, ...) 16:23:58 Sensor Voltage Measurement, Sensor Number = 1, V = 1.156348 V 16:23:58 Sensor Measurement, Sensor Number = 1, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 1.156348 (V, A, ohm, ...), Measured Value = 11.726070 (C, atm, Hz, ...) 16:23:58 Magnet Test Parameter Measurements P0 = MAIN, Value = 180.01 P1 = MAIN (V), Value = 7.11 P2 = Yoke, Value = 26.01 P3 = Bottom Coil, Value = 24.13 P4 = Top Coil, Value = 28.61 P5 = Water In, Value = 30.96 P6 = Water Out, Value = 28.66 P7 = Ambient, Value = 24.44 P8 = Supply Pres(psi), Value = 123.81 P9 = Return Pres(psi), Value = 11.73 16:23:58 Magnetic Field Measurement Measuring the magnet strength... 16:23:58 Integrated Voltage Measurement... 16:23:59 Magnet Current for MAIN, Imag = 180.0132 A Vtrans = 1.800185 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:24:07 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.006878753 Vs 16:24:15 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.006894926 Vs 16:24:16 Magnet Current for MAIN, Imag = 180.0128 A Vtrans = 1.800181 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:24:23 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.006883346 Vs 16:24:31 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.006895204 Vs 16:24:32 Magnet Current for MAIN, Imag = 180.0127 A Vtrans = 1.800180 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:24:39 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.006878840 Vs 16:24:47 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.006895257 Vs 16:24:47 Magnet Current for MAIN, Imag = 180.0127 A Vtrans = 1.800180 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:24:55 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.006881087 Vs 16:25:02 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.006899430 Vs 16:25:02 Integrated Voltage Average and RMS of Measurements X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.006888355 +- 0.000001455 Vs 16:25:02 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0000000 m, dx = 0.0100000 m, Nturns = 10.0 VT = 0.006888355 +- 0.000001455 Vs BL = 0.06888355 +- 0.00001455 Tm 16:25:06 Position moved: 0.000000 m 16:25:06 BL vs X Fit Coefficients: Cn sigCn n (Tm/m^n) ----- --------+-------- 0 0.06888 0.00001 Compute Fit Residuals: X0 BL sigBL BLfit BL-BLfit (m) (Tm) (Tm) (Tm) (Tm) ----------- --------+-------- -------- -------- 0.0000000 0.06888 0.00001 0.06888 0.00000 16:25:06 Magnet Strength Calculation Harmonic Number = 1 Number Of X0 Positions = 1 X0 BL sigBL (m) (Tm) (Tm) -------- --------+-------- 0.0000 0.06888 0.00001 SL = 0.0688836 +- 0.0000146 Tm^2/m^n 16:25:06 Magnet Strength Measurement Harmonic Number = 1 SL = 0.0688836 +- 0.0000146 Tm^2/m^n 16:25:09 Position moved: 0.000000 m 16:25:09 Magnet system MAIN current ramping to 165.0000 A... 16:25:21 Magnet Current for MAIN, Imag = 164.9770 A Vtrans = 1.649824 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:25:22 MAIN (V) Voltage Measurement, V = 6.511431 V 16:25:25 Temperature Measurement, Probe Number = 0, T = 26.012000 C 16:25:26 Temperature Measurement, Probe Number = 1, T = 24.159000 C 16:25:26 Temperature Measurement, Probe Number = 2, T = 28.601000 C 16:25:27 Temperature Measurement, Probe Number = 3, T = 30.932000 C 16:25:28 Temperature Measurement, Probe Number = 4, T = 28.609000 C 16:25:28 Temperature Measurement, Probe Number = 5, T = 24.475000 C 16:25:29 Sensor Voltage Measurement, Sensor Number = 0, V = 2.637182 V 16:25:29 Sensor Measurement, Sensor Number = 0, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 2.637182 (V, A, ohm, ...), Measured Value = 122.788612 (C, atm, Hz, ...) 16:25:29 Sensor Voltage Measurement, Sensor Number = 1, V = 1.156544 V 16:25:29 Sensor Measurement, Sensor Number = 1, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 1.156544 (V, A, ohm, ...), Measured Value = 11.740770 (C, atm, Hz, ...) 16:25:29 Magnet Test Parameter Measurements P0 = MAIN, Value = 164.98 P1 = MAIN (V), Value = 6.51 P2 = Yoke, Value = 26.01 P3 = Bottom Coil, Value = 24.16 P4 = Top Coil, Value = 28.60 P5 = Water In, Value = 30.93 P6 = Water Out, Value = 28.61 P7 = Ambient, Value = 24.48 P8 = Supply Pres(psi), Value = 122.79 P9 = Return Pres(psi), Value = 11.74 16:25:29 Magnetic Field Measurement Measuring the magnet strength... 16:25:29 Integrated Voltage Measurement... 16:25:30 Magnet Current for MAIN, Imag = 164.9768 A Vtrans = 1.649821 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:25:39 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.006315269 Vs 16:25:46 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.006327415 Vs 16:25:47 Magnet Current for MAIN, Imag = 164.9772 A Vtrans = 1.649825 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:25:55 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.006314805 Vs 16:26:02 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.006329157 Vs 16:26:03 Magnet Current for MAIN, Imag = 164.9769 A Vtrans = 1.649823 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:26:11 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.006312650 Vs 16:26:18 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.006333091 Vs 16:26:19 Magnet Current for MAIN, Imag = 164.9773 A Vtrans = 1.649826 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:26:27 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.006316131 Vs 16:26:34 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.006331043 Vs 16:26:34 Integrated Voltage Average and RMS of Measurements X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.006322445 +- 0.000000854 Vs 16:26:34 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0000000 m, dx = 0.0100000 m, Nturns = 10.0 VT = 0.006322445 +- 0.000000854 Vs BL = 0.06322445 +- 0.00000854 Tm 16:26:38 Position moved: 0.000000 m 16:26:38 BL vs X Fit Coefficients: Cn sigCn n (Tm/m^n) ----- --------+-------- 0 0.06322 0.00001 Compute Fit Residuals: X0 BL sigBL BLfit BL-BLfit (m) (Tm) (Tm) (Tm) (Tm) ----------- --------+-------- -------- -------- 0.0000000 0.06322 0.00001 0.06322 0.00000 16:26:38 Magnet Strength Calculation Harmonic Number = 1 Number Of X0 Positions = 1 X0 BL sigBL (m) (Tm) (Tm) -------- --------+-------- 0.0000 0.06322 0.00001 SL = 0.0632244 +- 0.0000085 Tm^2/m^n 16:26:38 Magnet Strength Measurement Harmonic Number = 1 SL = 0.0632244 +- 0.0000085 Tm^2/m^n 16:26:41 Position moved: 0.000000 m 16:26:41 Magnet system MAIN current ramping to 150.0000 A... 16:26:53 Magnet Current for MAIN, Imag = 150.0260 A Vtrans = 1.500313 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:26:54 MAIN (V) Voltage Measurement, V = 5.916150 V 16:26:57 Temperature Measurement, Probe Number = 0, T = 26.018000 C 16:26:58 Temperature Measurement, Probe Number = 1, T = 24.155000 C 16:26:58 Temperature Measurement, Probe Number = 2, T = 28.421000 C 16:26:59 Temperature Measurement, Probe Number = 3, T = 30.673000 C 16:27:00 Temperature Measurement, Probe Number = 4, T = 28.645000 C 16:27:00 Temperature Measurement, Probe Number = 5, T = 24.477000 C 16:27:01 Sensor Voltage Measurement, Sensor Number = 0, V = 2.653243 V 16:27:01 Sensor Measurement, Sensor Number = 0, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 2.653243 (V, A, ohm, ...), Measured Value = 123.993233 (C, atm, Hz, ...) 16:27:01 Sensor Voltage Measurement, Sensor Number = 1, V = 1.156410 V 16:27:01 Sensor Measurement, Sensor Number = 1, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 1.156410 (V, A, ohm, ...), Measured Value = 11.730750 (C, atm, Hz, ...) 16:27:01 Magnet Test Parameter Measurements P0 = MAIN, Value = 150.03 P1 = MAIN (V), Value = 5.92 P2 = Yoke, Value = 26.02 P3 = Bottom Coil, Value = 24.16 P4 = Top Coil, Value = 28.42 P5 = Water In, Value = 30.67 P6 = Water Out, Value = 28.64 P7 = Ambient, Value = 24.48 P8 = Supply Pres(psi), Value = 123.99 P9 = Return Pres(psi), Value = 11.73 16:27:01 Magnetic Field Measurement Measuring the magnet strength... 16:27:01 Integrated Voltage Measurement... 16:27:02 Magnet Current for MAIN, Imag = 150.0262 A Vtrans = 1.500316 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:27:11 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.005747874 Vs 16:27:18 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.005760638 Vs 16:27:19 Magnet Current for MAIN, Imag = 150.0262 A Vtrans = 1.500316 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:27:27 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.005743119 Vs 16:27:34 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.005763005 Vs 16:27:35 Magnet Current for MAIN, Imag = 150.0259 A Vtrans = 1.500312 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:27:43 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.005748453 Vs 16:27:50 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.005765787 Vs 16:27:51 Magnet Current for MAIN, Imag = 150.0262 A Vtrans = 1.500316 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:27:59 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.005747657 Vs 16:28:06 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.005766797 Vs 16:28:06 Integrated Voltage Average and RMS of Measurements X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.005755416 +- 0.000001808 Vs 16:28:06 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0000000 m, dx = 0.0100000 m, Nturns = 10.0 VT = 0.005755416 +- 0.000001808 Vs BL = 0.05755416 +- 0.00001808 Tm 16:28:10 Position moved: 0.000000 m 16:28:10 BL vs X Fit Coefficients: Cn sigCn n (Tm/m^n) ----- --------+-------- 0 0.05755 0.00002 Compute Fit Residuals: X0 BL sigBL BLfit BL-BLfit (m) (Tm) (Tm) (Tm) (Tm) ----------- --------+-------- -------- -------- 0.0000000 0.05755 0.00002 0.05755 0.00000 16:28:10 Magnet Strength Calculation Harmonic Number = 1 Number Of X0 Positions = 1 X0 BL sigBL (m) (Tm) (Tm) -------- --------+-------- 0.0000 0.05755 0.00002 SL = 0.0575542 +- 0.0000181 Tm^2/m^n 16:28:10 Magnet Strength Measurement Harmonic Number = 1 SL = 0.0575542 +- 0.0000181 Tm^2/m^n Remeasure 16:28:10 Magnetic Field Measurement Measuring the magnet strength... 16:28:10 Integrated Voltage Measurement... 16:28:11 Magnet Current for MAIN, Imag = 150.0257 A Vtrans = 1.500311 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:28:20 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.005746841 Vs 16:28:27 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.005762436 Vs 16:28:28 Magnet Current for MAIN, Imag = 150.0260 A Vtrans = 1.500313 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:28:36 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.005745600 Vs 16:28:43 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.005761744 Vs 16:28:44 Magnet Current for MAIN, Imag = 150.0260 A Vtrans = 1.500313 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:28:52 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.005746217 Vs 16:28:59 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.005763352 Vs 16:29:00 Magnet Current for MAIN, Imag = 150.0260 A Vtrans = 1.500313 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:29:08 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.005746359 Vs 16:29:15 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.005764586 Vs 16:29:15 Integrated Voltage Average and RMS of Measurements X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.005754642 +- 0.000000642 Vs 16:29:15 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0000000 m, dx = 0.0100000 m, Nturns = 10.0 VT = 0.005754642 +- 0.000000642 Vs BL = 0.05754642 +- 0.00000642 Tm 16:29:19 Position moved: 0.000000 m 16:29:19 BL vs X Fit Coefficients: Cn sigCn n (Tm/m^n) ----- --------+-------- 0 0.05755 0.00001 Compute Fit Residuals: X0 BL sigBL BLfit BL-BLfit (m) (Tm) (Tm) (Tm) (Tm) ----------- --------+-------- -------- -------- 0.0000000 0.05755 0.00001 0.05755 0.00000 16:29:19 Magnet Strength Calculation Harmonic Number = 1 Number Of X0 Positions = 1 X0 BL sigBL (m) (Tm) (Tm) -------- --------+-------- 0.0000 0.05755 0.00001 SL = 0.0575464 +- 0.0000064 Tm^2/m^n 16:29:19 Magnet Strength Measurement Harmonic Number = 1 SL = 0.0575464 +- 0.0000064 Tm^2/m^n 16:29:22 Position moved: 0.000000 m 16:29:22 Magnet system MAIN current ramping to 135.0000 A... 16:29:34 Magnet Current for MAIN, Imag = 134.9905 A Vtrans = 1.349959 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:29:35 MAIN (V) Voltage Measurement, V = 5.318885 V 16:29:38 Temperature Measurement, Probe Number = 0, T = 26.024000 C 16:29:39 Temperature Measurement, Probe Number = 1, T = 24.183000 C 16:29:40 Temperature Measurement, Probe Number = 2, T = 27.992000 C 16:29:40 Temperature Measurement, Probe Number = 3, T = 30.300000 C 16:29:41 Temperature Measurement, Probe Number = 4, T = 28.671000 C 16:29:41 Temperature Measurement, Probe Number = 5, T = 24.494000 C 16:29:42 Sensor Voltage Measurement, Sensor Number = 0, V = 2.645267 V 16:29:42 Sensor Measurement, Sensor Number = 0, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 2.645267 (V, A, ohm, ...), Measured Value = 123.394995 (C, atm, Hz, ...) 16:29:42 Sensor Voltage Measurement, Sensor Number = 1, V = 1.156040 V 16:29:42 Sensor Measurement, Sensor Number = 1, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 1.156040 (V, A, ohm, ...), Measured Value = 11.703030 (C, atm, Hz, ...) 16:29:42 Magnet Test Parameter Measurements P0 = MAIN, Value = 134.99 P1 = MAIN (V), Value = 5.32 P2 = Yoke, Value = 26.02 P3 = Bottom Coil, Value = 24.18 P4 = Top Coil, Value = 27.99 P5 = Water In, Value = 30.30 P6 = Water Out, Value = 28.67 P7 = Ambient, Value = 24.49 P8 = Supply Pres(psi), Value = 123.39 P9 = Return Pres(psi), Value = 11.70 16:29:42 Magnetic Field Measurement Measuring the magnet strength... 16:29:42 Integrated Voltage Measurement... 16:29:43 Magnet Current for MAIN, Imag = 134.9906 A Vtrans = 1.349960 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:29:52 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.005176005 Vs 16:29:59 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.005189151 Vs 16:30:00 Magnet Current for MAIN, Imag = 134.9908 A Vtrans = 1.349961 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:30:08 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.005176854 Vs 16:30:16 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.005193402 Vs 16:30:16 Magnet Current for MAIN, Imag = 134.9906 A Vtrans = 1.349960 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:30:24 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.005175505 Vs 16:30:31 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.005195478 Vs 16:30:32 Magnet Current for MAIN, Imag = 134.9911 A Vtrans = 1.349965 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:30:40 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.005173829 Vs 16:30:47 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.005191191 Vs 16:30:47 Integrated Voltage Average and RMS of Measurements X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.005183927 +- 0.000001389 Vs 16:30:47 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0000000 m, dx = 0.0100000 m, Nturns = 10.0 VT = 0.005183927 +- 0.000001389 Vs BL = 0.05183927 +- 0.00001389 Tm 16:30:51 Position moved: 0.000000 m 16:30:51 BL vs X Fit Coefficients: Cn sigCn n (Tm/m^n) ----- --------+-------- 0 0.05184 0.00001 Compute Fit Residuals: X0 BL sigBL BLfit BL-BLfit (m) (Tm) (Tm) (Tm) (Tm) ----------- --------+-------- -------- -------- 0.0000000 0.05184 0.00001 0.05184 0.00000 16:30:51 Magnet Strength Calculation Harmonic Number = 1 Number Of X0 Positions = 1 X0 BL sigBL (m) (Tm) (Tm) -------- --------+-------- 0.0000 0.05184 0.00001 SL = 0.0518393 +- 0.0000139 Tm^2/m^n 16:30:51 Magnet Strength Measurement Harmonic Number = 1 SL = 0.0518393 +- 0.0000139 Tm^2/m^n Remeasure 16:30:51 Magnetic Field Measurement Measuring the magnet strength... 16:30:51 Integrated Voltage Measurement... 16:30:52 Magnet Current for MAIN, Imag = 134.9906 A Vtrans = 1.349960 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:31:01 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.005174820 Vs 16:31:08 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.005191909 Vs 16:31:09 Magnet Current for MAIN, Imag = 134.9910 A Vtrans = 1.349964 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:31:17 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.005175457 Vs 16:31:24 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.005192339 Vs 16:31:25 Magnet Current for MAIN, Imag = 134.9905 A Vtrans = 1.349959 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:31:33 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.005173793 Vs 16:31:40 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.005191294 Vs 16:31:41 Magnet Current for MAIN, Imag = 134.9910 A Vtrans = 1.349964 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:31:49 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.005174586 Vs 16:31:56 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.005195306 Vs 16:31:56 Integrated Voltage Average and RMS of Measurements X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.005183688 +- 0.000000872 Vs 16:31:56 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0000000 m, dx = 0.0100000 m, Nturns = 10.0 VT = 0.005183688 +- 0.000000872 Vs BL = 0.05183688 +- 0.00000872 Tm 16:32:00 Position moved: 0.000000 m 16:32:00 BL vs X Fit Coefficients: Cn sigCn n (Tm/m^n) ----- --------+-------- 0 0.05184 0.00001 Compute Fit Residuals: X0 BL sigBL BLfit BL-BLfit (m) (Tm) (Tm) (Tm) (Tm) ----------- --------+-------- -------- -------- 0.0000000 0.05184 0.00001 0.05184 0.00000 16:32:00 Magnet Strength Calculation Harmonic Number = 1 Number Of X0 Positions = 1 X0 BL sigBL (m) (Tm) (Tm) -------- --------+-------- 0.0000 0.05184 0.00001 SL = 0.0518369 +- 0.0000087 Tm^2/m^n 16:32:00 Magnet Strength Measurement Harmonic Number = 1 SL = 0.0518369 +- 0.0000087 Tm^2/m^n 16:32:03 Position moved: 0.000000 m 16:32:03 Magnet system MAIN current ramping to 120.0000 A... 16:32:15 Magnet Current for MAIN, Imag = 120.0366 A Vtrans = 1.200419 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:32:16 MAIN (V) Voltage Measurement, V = 4.726232 V 16:32:19 Temperature Measurement, Probe Number = 0, T = 26.046000 C 16:32:20 Temperature Measurement, Probe Number = 1, T = 24.239000 C 16:32:20 Temperature Measurement, Probe Number = 2, T = 27.903000 C 16:32:21 Temperature Measurement, Probe Number = 3, T = 29.995000 C 16:32:21 Temperature Measurement, Probe Number = 4, T = 28.726000 C 16:32:22 Temperature Measurement, Probe Number = 5, T = 24.493000 C 16:32:23 Sensor Voltage Measurement, Sensor Number = 0, V = 2.656431 V 16:32:23 Sensor Measurement, Sensor Number = 0, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 2.656431 (V, A, ohm, ...), Measured Value = 124.232318 (C, atm, Hz, ...) 16:32:23 Sensor Voltage Measurement, Sensor Number = 1, V = 1.155310 V 16:32:23 Sensor Measurement, Sensor Number = 1, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 1.155310 (V, A, ohm, ...), Measured Value = 11.648243 (C, atm, Hz, ...) 16:32:23 Magnet Test Parameter Measurements P0 = MAIN, Value = 120.04 P1 = MAIN (V), Value = 4.73 P2 = Yoke, Value = 26.05 P3 = Bottom Coil, Value = 24.24 P4 = Top Coil, Value = 27.90 P5 = Water In, Value = 30.00 P6 = Water Out, Value = 28.73 P7 = Ambient, Value = 24.49 P8 = Supply Pres(psi), Value = 124.23 P9 = Return Pres(psi), Value = 11.65 16:32:23 Magnetic Field Measurement Measuring the magnet strength... 16:32:23 Integrated Voltage Measurement... 16:32:24 Magnet Current for MAIN, Imag = 120.0367 A Vtrans = 1.200420 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:32:33 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.004601387 Vs 16:32:40 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.004622250 Vs 16:32:41 Magnet Current for MAIN, Imag = 120.0365 A Vtrans = 1.200418 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:32:48 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.004604754 Vs 16:32:56 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.004623103 Vs 16:32:57 Magnet Current for MAIN, Imag = 120.0366 A Vtrans = 1.200419 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:33:04 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.004603361 Vs 16:33:12 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.004622233 Vs 16:33:13 Magnet Current for MAIN, Imag = 120.0366 A Vtrans = 1.200419 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:33:20 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.004604116 Vs 16:33:28 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.004619010 Vs 16:33:28 Integrated Voltage Average and RMS of Measurements X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.004612527 +- 0.000000931 Vs 16:33:28 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0000000 m, dx = 0.0100000 m, Nturns = 10.0 VT = 0.004612527 +- 0.000000931 Vs BL = 0.04612527 +- 0.00000931 Tm 16:33:32 Position moved: 0.000000 m 16:33:32 BL vs X Fit Coefficients: Cn sigCn n (Tm/m^n) ----- --------+-------- 0 0.04613 0.00001 Compute Fit Residuals: X0 BL sigBL BLfit BL-BLfit (m) (Tm) (Tm) (Tm) (Tm) ----------- --------+-------- -------- -------- 0.0000000 0.04613 0.00001 0.04613 0.00000 16:33:32 Magnet Strength Calculation Harmonic Number = 1 Number Of X0 Positions = 1 X0 BL sigBL (m) (Tm) (Tm) -------- --------+-------- 0.0000 0.04613 0.00001 SL = 0.0461253 +- 0.0000093 Tm^2/m^n 16:33:32 Magnet Strength Measurement Harmonic Number = 1 SL = 0.0461253 +- 0.0000093 Tm^2/m^n 16:33:35 Position moved: 0.000000 m 16:33:35 Magnet system MAIN current ramping to 105.0000 A... 16:33:47 Magnet Current for MAIN, Imag = 105.0184 A Vtrans = 1.050237 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:33:47 MAIN (V) Voltage Measurement, V = 4.132313 V 16:33:51 Temperature Measurement, Probe Number = 0, T = 26.057000 C 16:33:52 Temperature Measurement, Probe Number = 1, T = 24.224000 C 16:33:52 Temperature Measurement, Probe Number = 2, T = 27.836000 C 16:33:53 Temperature Measurement, Probe Number = 3, T = 29.767000 C 16:33:53 Temperature Measurement, Probe Number = 4, T = 28.678000 C 16:33:54 Temperature Measurement, Probe Number = 5, T = 24.489000 C 16:33:55 Sensor Voltage Measurement, Sensor Number = 0, V = 2.651532 V 16:33:55 Sensor Measurement, Sensor Number = 0, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 2.651532 (V, A, ohm, ...), Measured Value = 123.864930 (C, atm, Hz, ...) 16:33:55 Sensor Voltage Measurement, Sensor Number = 1, V = 1.156790 V 16:33:55 Sensor Measurement, Sensor Number = 1, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 1.156790 (V, A, ohm, ...), Measured Value = 11.759220 (C, atm, Hz, ...) 16:33:55 Magnet Test Parameter Measurements P0 = MAIN, Value = 105.02 P1 = MAIN (V), Value = 4.13 P2 = Yoke, Value = 26.06 P3 = Bottom Coil, Value = 24.22 P4 = Top Coil, Value = 27.84 P5 = Water In, Value = 29.77 P6 = Water Out, Value = 28.68 P7 = Ambient, Value = 24.49 P8 = Supply Pres(psi), Value = 123.86 P9 = Return Pres(psi), Value = 11.76 16:33:55 Magnetic Field Measurement Measuring the magnet strength... 16:33:55 Integrated Voltage Measurement... 16:33:56 Magnet Current for MAIN, Imag = 105.0180 A Vtrans = 1.050233 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:34:05 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.004029020 Vs 16:34:12 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.004048916 Vs 16:34:13 Magnet Current for MAIN, Imag = 105.0186 A Vtrans = 1.050240 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:34:21 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.004028540 Vs 16:34:28 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.004043781 Vs 16:34:29 Magnet Current for MAIN, Imag = 105.0184 A Vtrans = 1.050237 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:34:37 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.004030169 Vs 16:34:44 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.004048754 Vs 16:34:45 Magnet Current for MAIN, Imag = 105.0185 A Vtrans = 1.050238 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:34:53 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.004028711 Vs 16:35:00 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.004047994 Vs 16:35:00 Integrated Voltage Average and RMS of Measurements X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.004038235 +- 0.000001261 Vs 16:35:00 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0000000 m, dx = 0.0100000 m, Nturns = 10.0 VT = 0.004038235 +- 0.000001261 Vs BL = 0.04038235 +- 0.00001261 Tm 16:35:04 Position moved: 0.000000 m 16:35:04 BL vs X Fit Coefficients: Cn sigCn n (Tm/m^n) ----- --------+-------- 0 0.04038 0.00001 Compute Fit Residuals: X0 BL sigBL BLfit BL-BLfit (m) (Tm) (Tm) (Tm) (Tm) ----------- --------+-------- -------- -------- 0.0000000 0.04038 0.00001 0.04038 0.00000 16:35:04 Magnet Strength Calculation Harmonic Number = 1 Number Of X0 Positions = 1 X0 BL sigBL (m) (Tm) (Tm) -------- --------+-------- 0.0000 0.04038 0.00001 SL = 0.0403824 +- 0.0000126 Tm^2/m^n 16:35:04 Magnet Strength Measurement Harmonic Number = 1 SL = 0.0403824 +- 0.0000126 Tm^2/m^n Remeasure 16:35:04 Magnetic Field Measurement Measuring the magnet strength... 16:35:04 Integrated Voltage Measurement... 16:35:05 Magnet Current for MAIN, Imag = 105.0186 A Vtrans = 1.050240 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:35:14 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.004029811 Vs 16:35:21 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.004047467 Vs 16:35:22 Magnet Current for MAIN, Imag = 105.0181 A Vtrans = 1.050235 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:35:30 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.004030679 Vs 16:35:37 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.004048248 Vs 16:35:38 Magnet Current for MAIN, Imag = 105.0182 A Vtrans = 1.050236 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:35:46 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.004031454 Vs 16:35:53 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.004048714 Vs 16:35:54 Magnet Current for MAIN, Imag = 105.0184 A Vtrans = 1.050237 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:36:02 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.004030121 Vs 16:36:09 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.004046440 Vs 16:36:09 Integrated Voltage Average and RMS of Measurements X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.004039117 +- 0.000000704 Vs 16:36:09 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0000000 m, dx = 0.0100000 m, Nturns = 10.0 VT = 0.004039117 +- 0.000000704 Vs BL = 0.04039117 +- 0.00000704 Tm 16:36:13 Position moved: 0.000000 m 16:36:13 BL vs X Fit Coefficients: Cn sigCn n (Tm/m^n) ----- --------+-------- 0 0.04039 0.00001 Compute Fit Residuals: X0 BL sigBL BLfit BL-BLfit (m) (Tm) (Tm) (Tm) (Tm) ----------- --------+-------- -------- -------- 0.0000000 0.04039 0.00001 0.04039 0.00000 16:36:13 Magnet Strength Calculation Harmonic Number = 1 Number Of X0 Positions = 1 X0 BL sigBL (m) (Tm) (Tm) -------- --------+-------- 0.0000 0.04039 0.00001 SL = 0.0403912 +- 0.0000070 Tm^2/m^n 16:36:13 Magnet Strength Measurement Harmonic Number = 1 SL = 0.0403912 +- 0.0000070 Tm^2/m^n 16:36:16 Position moved: 0.000000 m 16:36:16 Magnet system MAIN current ramping to 90.0000 A... 16:36:28 Magnet Current for MAIN, Imag = 89.9904 A Vtrans = 0.899957 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:36:29 MAIN (V) Voltage Measurement, V = 3.539085 V 16:36:32 Temperature Measurement, Probe Number = 0, T = 26.057000 C 16:36:33 Temperature Measurement, Probe Number = 1, T = 24.184000 C 16:36:33 Temperature Measurement, Probe Number = 2, T = 27.798000 C 16:36:34 Temperature Measurement, Probe Number = 3, T = 29.467000 C 16:36:35 Temperature Measurement, Probe Number = 4, T = 28.678000 C 16:36:35 Temperature Measurement, Probe Number = 5, T = 24.537000 C 16:36:36 Sensor Voltage Measurement, Sensor Number = 0, V = 2.654349 V 16:36:36 Sensor Measurement, Sensor Number = 0, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 2.654349 (V, A, ohm, ...), Measured Value = 124.076205 (C, atm, Hz, ...) 16:36:36 Sensor Voltage Measurement, Sensor Number = 1, V = 1.153800 V 16:36:36 Sensor Measurement, Sensor Number = 1, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 1.153800 (V, A, ohm, ...), Measured Value = 11.535022 (C, atm, Hz, ...) 16:36:36 Magnet Test Parameter Measurements P0 = MAIN, Value = 89.99 P1 = MAIN (V), Value = 3.54 P2 = Yoke, Value = 26.06 P3 = Bottom Coil, Value = 24.18 P4 = Top Coil, Value = 27.80 P5 = Water In, Value = 29.47 P6 = Water Out, Value = 28.68 P7 = Ambient, Value = 24.54 P8 = Supply Pres(psi), Value = 124.08 P9 = Return Pres(psi), Value = 11.54 16:36:36 Magnetic Field Measurement Measuring the magnet strength... 16:36:36 Integrated Voltage Measurement... 16:36:37 Magnet Current for MAIN, Imag = 89.9904 A Vtrans = 0.899958 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:36:46 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.003456433 Vs 16:36:53 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.003471227 Vs 16:36:54 Magnet Current for MAIN, Imag = 89.9903 A Vtrans = 0.899956 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:37:02 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.003452992 Vs 16:37:09 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.003472726 Vs 16:37:10 Magnet Current for MAIN, Imag = 89.9902 A Vtrans = 0.899955 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:37:18 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.003454766 Vs 16:37:25 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.003470643 Vs 16:37:26 Magnet Current for MAIN, Imag = 89.9905 A Vtrans = 0.899958 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:37:34 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.003456111 Vs 16:37:41 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.003472668 Vs 16:37:41 Integrated Voltage Average and RMS of Measurements X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.003463446 +- 0.000000695 Vs 16:37:41 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0000000 m, dx = 0.0100000 m, Nturns = 10.0 VT = 0.003463446 +- 0.000000695 Vs BL = 0.03463446 +- 0.00000695 Tm 16:37:45 Position moved: 0.000000 m 16:37:45 BL vs X Fit Coefficients: Cn sigCn n (Tm/m^n) ----- --------+-------- 0 0.03463 0.00001 Compute Fit Residuals: X0 BL sigBL BLfit BL-BLfit (m) (Tm) (Tm) (Tm) (Tm) ----------- --------+-------- -------- -------- 0.0000000 0.03463 0.00001 0.03463 0.00000 16:37:45 Magnet Strength Calculation Harmonic Number = 1 Number Of X0 Positions = 1 X0 BL sigBL (m) (Tm) (Tm) -------- --------+-------- 0.0000 0.03463 0.00001 SL = 0.0346345 +- 0.0000069 Tm^2/m^n 16:37:45 Magnet Strength Measurement Harmonic Number = 1 SL = 0.0346345 +- 0.0000069 Tm^2/m^n 16:37:48 Position moved: 0.000000 m 16:37:48 Magnet system MAIN current ramping to 75.0000 A... 16:38:00 Magnet Current for MAIN, Imag = 75.0360 A Vtrans = 0.750413 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:38:01 MAIN (V) Voltage Measurement, V = 2.949624 V 16:38:04 Temperature Measurement, Probe Number = 0, T = 26.060000 C 16:38:05 Temperature Measurement, Probe Number = 1, T = 24.208000 C 16:38:06 Temperature Measurement, Probe Number = 2, T = 27.666000 C 16:38:06 Temperature Measurement, Probe Number = 3, T = 29.301000 C 16:38:07 Temperature Measurement, Probe Number = 4, T = 28.706000 C 16:38:07 Temperature Measurement, Probe Number = 5, T = 24.525000 C 16:38:08 Sensor Voltage Measurement, Sensor Number = 0, V = 2.644279 V 16:38:08 Sensor Measurement, Sensor Number = 0, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 2.644279 (V, A, ohm, ...), Measured Value = 123.320918 (C, atm, Hz, ...) 16:38:08 Sensor Voltage Measurement, Sensor Number = 1, V = 1.156004 V 16:38:08 Sensor Measurement, Sensor Number = 1, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 1.156004 (V, A, ohm, ...), Measured Value = 11.700315 (C, atm, Hz, ...) 16:38:08 Magnet Test Parameter Measurements P0 = MAIN, Value = 75.04 P1 = MAIN (V), Value = 2.95 P2 = Yoke, Value = 26.06 P3 = Bottom Coil, Value = 24.21 P4 = Top Coil, Value = 27.67 P5 = Water In, Value = 29.30 P6 = Water Out, Value = 28.71 P7 = Ambient, Value = 24.52 P8 = Supply Pres(psi), Value = 123.32 P9 = Return Pres(psi), Value = 11.70 16:38:08 Magnetic Field Measurement Measuring the magnet strength... 16:38:08 Integrated Voltage Measurement... 16:38:09 Magnet Current for MAIN, Imag = 75.0362 A Vtrans = 0.750416 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:38:18 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.002880090 Vs 16:38:25 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.002899031 Vs 16:38:26 Magnet Current for MAIN, Imag = 75.0361 A Vtrans = 0.750414 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:38:34 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.002881477 Vs 16:38:41 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.002896813 Vs 16:38:42 Magnet Current for MAIN, Imag = 75.0362 A Vtrans = 0.750415 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:38:50 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.002881444 Vs 16:38:57 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.002896438 Vs 16:38:58 Magnet Current for MAIN, Imag = 75.0362 A Vtrans = 0.750415 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:39:06 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.002881395 Vs 16:39:14 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.002898249 Vs 16:39:14 Integrated Voltage Average and RMS of Measurements X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.002889367 +- 0.000000345 Vs 16:39:14 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0000000 m, dx = 0.0100000 m, Nturns = 10.0 VT = 0.002889367 +- 0.000000345 Vs BL = 0.02889367 +- 0.00000345 Tm 16:39:17 Position moved: 0.000000 m 16:39:17 BL vs X Fit Coefficients: Cn sigCn n (Tm/m^n) ----- --------+-------- 0 0.02889 0.00000 Compute Fit Residuals: X0 BL sigBL BLfit BL-BLfit (m) (Tm) (Tm) (Tm) (Tm) ----------- --------+-------- -------- -------- 0.0000000 0.02889 0.00000 0.02889 0.00000 16:39:17 Magnet Strength Calculation Harmonic Number = 1 Number Of X0 Positions = 1 X0 BL sigBL (m) (Tm) (Tm) -------- --------+-------- 0.0000 0.02889 0.00000 SL = 0.0288937 +- 0.0000034 Tm^2/m^n 16:39:17 Magnet Strength Measurement Harmonic Number = 1 SL = 0.0288937 +- 0.0000034 Tm^2/m^n 16:39:20 Position moved: 0.000000 m 16:39:20 Magnet system MAIN current ramping to 60.0000 A... 16:39:32 Magnet Current for MAIN, Imag = 60.0068 A Vtrans = 0.600122 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:39:33 MAIN (V) Voltage Measurement, V = 2.357938 V 16:39:36 Temperature Measurement, Probe Number = 0, T = 26.068000 C 16:39:37 Temperature Measurement, Probe Number = 1, T = 24.237000 C 16:39:38 Temperature Measurement, Probe Number = 2, T = 27.576000 C 16:39:38 Temperature Measurement, Probe Number = 3, T = 29.136000 C 16:39:39 Temperature Measurement, Probe Number = 4, T = 28.699000 C 16:39:39 Temperature Measurement, Probe Number = 5, T = 24.518000 C 16:39:40 Sensor Voltage Measurement, Sensor Number = 0, V = 2.643406 V 16:39:40 Sensor Measurement, Sensor Number = 0, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 2.643406 (V, A, ohm, ...), Measured Value = 123.255458 (C, atm, Hz, ...) 16:39:40 Sensor Voltage Measurement, Sensor Number = 1, V = 1.155668 V 16:39:41 Sensor Measurement, Sensor Number = 1, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 1.155668 (V, A, ohm, ...), Measured Value = 11.675123 (C, atm, Hz, ...) 16:39:41 Magnet Test Parameter Measurements P0 = MAIN, Value = 60.01 P1 = MAIN (V), Value = 2.36 P2 = Yoke, Value = 26.07 P3 = Bottom Coil, Value = 24.24 P4 = Top Coil, Value = 27.58 P5 = Water In, Value = 29.14 P6 = Water Out, Value = 28.70 P7 = Ambient, Value = 24.52 P8 = Supply Pres(psi), Value = 123.26 P9 = Return Pres(psi), Value = 11.68 16:39:41 Magnetic Field Measurement Measuring the magnet strength... 16:39:41 Integrated Voltage Measurement... 16:39:41 Magnet Current for MAIN, Imag = 60.0065 A Vtrans = 0.600119 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:39:50 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.002305793 Vs 16:39:57 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.002320302 Vs 16:39:58 Magnet Current for MAIN, Imag = 60.0064 A Vtrans = 0.600117 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:40:06 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.002307462 Vs 16:40:13 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.002318960 Vs 16:40:14 Magnet Current for MAIN, Imag = 60.0062 A Vtrans = 0.600116 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:40:22 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.002305282 Vs 16:40:29 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.002322653 Vs 16:40:30 Magnet Current for MAIN, Imag = 60.0065 A Vtrans = 0.600119 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:40:38 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.002304610 Vs 16:40:45 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.002321673 Vs 16:40:45 Integrated Voltage Average and RMS of Measurements X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.002313342 +- 0.000000366 Vs 16:40:45 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0000000 m, dx = 0.0100000 m, Nturns = 10.0 VT = 0.002313342 +- 0.000000366 Vs BL = 0.02313342 +- 0.00000366 Tm 16:40:49 Position moved: 0.000000 m 16:40:49 BL vs X Fit Coefficients: Cn sigCn n (Tm/m^n) ----- --------+-------- 0 0.02313 0.00000 Compute Fit Residuals: X0 BL sigBL BLfit BL-BLfit (m) (Tm) (Tm) (Tm) (Tm) ----------- --------+-------- -------- -------- 0.0000000 0.02313 0.00000 0.02313 0.00000 16:40:49 Magnet Strength Calculation Harmonic Number = 1 Number Of X0 Positions = 1 X0 BL sigBL (m) (Tm) (Tm) -------- --------+-------- 0.0000 0.02313 0.00000 SL = 0.0231334 +- 0.0000037 Tm^2/m^n 16:40:49 Magnet Strength Measurement Harmonic Number = 1 SL = 0.0231334 +- 0.0000037 Tm^2/m^n 16:40:52 Position moved: 0.000000 m 16:40:52 Magnet system MAIN current ramping to 45.0000 A... 16:41:04 Magnet Current for MAIN, Imag = 44.9811 A Vtrans = 0.449864 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:41:05 MAIN (V) Voltage Measurement, V = 1.767000 V 16:41:08 Temperature Measurement, Probe Number = 0, T = 26.065000 C 16:41:09 Temperature Measurement, Probe Number = 1, T = 24.218000 C 16:41:09 Temperature Measurement, Probe Number = 2, T = 27.600000 C 16:41:10 Temperature Measurement, Probe Number = 3, T = 28.989000 C 16:41:11 Temperature Measurement, Probe Number = 4, T = 28.717000 C 16:41:11 Temperature Measurement, Probe Number = 5, T = 24.474000 C 16:41:12 Sensor Voltage Measurement, Sensor Number = 0, V = 2.652842 V 16:41:12 Sensor Measurement, Sensor Number = 0, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 2.652842 (V, A, ohm, ...), Measured Value = 123.963173 (C, atm, Hz, ...) 16:41:12 Sensor Voltage Measurement, Sensor Number = 1, V = 1.155399 V 16:41:12 Sensor Measurement, Sensor Number = 1, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 1.155399 (V, A, ohm, ...), Measured Value = 11.654895 (C, atm, Hz, ...) 16:41:12 Magnet Test Parameter Measurements P0 = MAIN, Value = 44.98 P1 = MAIN (V), Value = 1.77 P2 = Yoke, Value = 26.07 P3 = Bottom Coil, Value = 24.22 P4 = Top Coil, Value = 27.60 P5 = Water In, Value = 28.99 P6 = Water Out, Value = 28.72 P7 = Ambient, Value = 24.47 P8 = Supply Pres(psi), Value = 123.96 P9 = Return Pres(psi), Value = 11.65 16:41:12 Magnetic Field Measurement Measuring the magnet strength... 16:41:12 Integrated Voltage Measurement... 16:41:13 Magnet Current for MAIN, Imag = 44.9809 A Vtrans = 0.449863 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:41:22 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.001728868 Vs 16:41:29 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.001743510 Vs 16:41:30 Magnet Current for MAIN, Imag = 44.9811 A Vtrans = 0.449864 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:41:38 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.001729642 Vs 16:41:45 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.001743849 Vs 16:41:46 Magnet Current for MAIN, Imag = 44.9810 A Vtrans = 0.449864 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:41:54 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.001730144 Vs 16:42:01 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.001743002 Vs 16:42:02 Magnet Current for MAIN, Imag = 44.9811 A Vtrans = 0.449864 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:42:10 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.001726564 Vs 16:42:17 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.001745169 Vs 16:42:17 Integrated Voltage Average and RMS of Measurements X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.001736343 +- 0.000000341 Vs 16:42:17 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0000000 m, dx = 0.0100000 m, Nturns = 10.0 VT = 0.001736343 +- 0.000000341 Vs BL = 0.01736343 +- 0.00000341 Tm 16:42:21 Position moved: 0.000000 m 16:42:21 BL vs X Fit Coefficients: Cn sigCn n (Tm/m^n) ----- --------+-------- 0 0.01736 0.00000 Compute Fit Residuals: X0 BL sigBL BLfit BL-BLfit (m) (Tm) (Tm) (Tm) (Tm) ----------- --------+-------- -------- -------- 0.0000000 0.01736 0.00000 0.01736 0.00000 16:42:21 Magnet Strength Calculation Harmonic Number = 1 Number Of X0 Positions = 1 X0 BL sigBL (m) (Tm) (Tm) -------- --------+-------- 0.0000 0.01736 0.00000 SL = 0.0173634 +- 0.0000034 Tm^2/m^n 16:42:21 Magnet Strength Measurement Harmonic Number = 1 SL = 0.0173634 +- 0.0000034 Tm^2/m^n 16:42:24 Position moved: 0.000000 m 16:42:24 Magnet system MAIN current ramping to 30.0000 A... 16:42:36 Magnet Current for MAIN, Imag = 30.0286 A Vtrans = 0.300339 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:42:37 MAIN (V) Voltage Measurement, V = 1.179305 V 16:42:40 Temperature Measurement, Probe Number = 0, T = 26.070000 C 16:42:41 Temperature Measurement, Probe Number = 1, T = 24.170000 C 16:42:41 Temperature Measurement, Probe Number = 2, T = 27.477000 C 16:42:42 Temperature Measurement, Probe Number = 3, T = 28.857000 C 16:42:43 Temperature Measurement, Probe Number = 4, T = 28.737000 C 16:42:43 Temperature Measurement, Probe Number = 5, T = 24.441000 C 16:42:44 Sensor Voltage Measurement, Sensor Number = 0, V = 2.637350 V 16:42:44 Sensor Measurement, Sensor Number = 0, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 2.637350 (V, A, ohm, ...), Measured Value = 122.801257 (C, atm, Hz, ...) 16:42:44 Sensor Voltage Measurement, Sensor Number = 1, V = 1.155074 V 16:42:44 Sensor Measurement, Sensor Number = 1, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 1.155074 (V, A, ohm, ...), Measured Value = 11.630543 (C, atm, Hz, ...) 16:42:44 Magnet Test Parameter Measurements P0 = MAIN, Value = 30.03 P1 = MAIN (V), Value = 1.18 P2 = Yoke, Value = 26.07 P3 = Bottom Coil, Value = 24.17 P4 = Top Coil, Value = 27.48 P5 = Water In, Value = 28.86 P6 = Water Out, Value = 28.74 P7 = Ambient, Value = 24.44 P8 = Supply Pres(psi), Value = 122.80 P9 = Return Pres(psi), Value = 11.63 16:42:44 Magnetic Field Measurement Measuring the magnet strength... 16:42:44 Integrated Voltage Measurement... 16:42:45 Magnet Current for MAIN, Imag = 30.0286 A Vtrans = 0.300339 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:42:54 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.001155219 Vs 16:43:01 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.001171145 Vs 16:43:02 Magnet Current for MAIN, Imag = 30.0288 A Vtrans = 0.300341 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:43:10 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.001155019 Vs 16:43:18 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.001172400 Vs 16:43:18 Magnet Current for MAIN, Imag = 30.0284 A Vtrans = 0.300338 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:43:26 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.001153417 Vs 16:43:34 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.001171338 Vs 16:43:34 Magnet Current for MAIN, Imag = 30.0285 A Vtrans = 0.300339 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:43:42 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.001153671 Vs 16:43:50 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.001171590 Vs 16:43:50 Integrated Voltage Average and RMS of Measurements X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.001162975 +- 0.000000514 Vs 16:43:50 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0000000 m, dx = 0.0100000 m, Nturns = 10.0 VT = 0.001162975 +- 0.000000514 Vs BL = 0.01162975 +- 0.00000514 Tm 16:43:53 Position moved: 0.000000 m 16:43:53 BL vs X Fit Coefficients: Cn sigCn n (Tm/m^n) ----- --------+-------- 0 0.01163 0.00001 Compute Fit Residuals: X0 BL sigBL BLfit BL-BLfit (m) (Tm) (Tm) (Tm) (Tm) ----------- --------+-------- -------- -------- 0.0000000 0.01163 0.00001 0.01163 0.00000 16:43:53 Magnet Strength Calculation Harmonic Number = 1 Number Of X0 Positions = 1 X0 BL sigBL (m) (Tm) (Tm) -------- --------+-------- 0.0000 0.01163 0.00001 SL = 0.0116298 +- 0.0000051 Tm^2/m^n 16:43:53 Magnet Strength Measurement Harmonic Number = 1 SL = 0.0116298 +- 0.0000051 Tm^2/m^n Remeasure 16:43:53 Magnetic Field Measurement Measuring the magnet strength... 16:43:53 Integrated Voltage Measurement... 16:43:54 Magnet Current for MAIN, Imag = 30.0285 A Vtrans = 0.300338 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:44:03 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.001151970 Vs 16:44:10 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.001170943 Vs 16:44:11 Magnet Current for MAIN, Imag = 30.0285 A Vtrans = 0.300339 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:44:19 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.001154425 Vs 16:44:26 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.001172254 Vs 16:44:27 Magnet Current for MAIN, Imag = 30.0284 A Vtrans = 0.300338 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:44:35 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.001152567 Vs 16:44:42 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.001173232 Vs 16:44:43 Magnet Current for MAIN, Imag = 30.0286 A Vtrans = 0.300340 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:44:50 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.001153234 Vs 16:44:58 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.001171699 Vs 16:44:58 Integrated Voltage Average and RMS of Measurements X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.001162541 +- 0.000000698 Vs 16:44:58 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0000000 m, dx = 0.0100000 m, Nturns = 10.0 VT = 0.001162541 +- 0.000000698 Vs BL = 0.01162541 +- 0.00000698 Tm 16:45:02 Position moved: 0.000000 m 16:45:02 BL vs X Fit Coefficients: Cn sigCn n (Tm/m^n) ----- --------+-------- 0 0.01163 0.00001 Compute Fit Residuals: X0 BL sigBL BLfit BL-BLfit (m) (Tm) (Tm) (Tm) (Tm) ----------- --------+-------- -------- -------- 0.0000000 0.01163 0.00001 0.01163 0.00000 16:45:02 Magnet Strength Calculation Harmonic Number = 1 Number Of X0 Positions = 1 X0 BL sigBL (m) (Tm) (Tm) -------- --------+-------- 0.0000 0.01163 0.00001 SL = 0.0116254 +- 0.0000070 Tm^2/m^n 16:45:02 Magnet Strength Measurement Harmonic Number = 1 SL = 0.0116254 +- 0.0000070 Tm^2/m^n Remeasure 16:45:02 Magnetic Field Measurement Measuring the magnet strength... 16:45:02 Integrated Voltage Measurement... 16:45:02 Magnet Current for MAIN, Imag = 30.0284 A Vtrans = 0.300337 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:45:11 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.001153712 Vs 16:45:19 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.001171822 Vs 16:45:19 Magnet Current for MAIN, Imag = 30.0285 A Vtrans = 0.300338 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:45:27 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.001154481 Vs 16:45:35 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.001171205 Vs 16:45:35 Magnet Current for MAIN, Imag = 30.0286 A Vtrans = 0.300339 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:45:43 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.001155122 Vs 16:45:50 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.001170961 Vs 16:45:51 Magnet Current for MAIN, Imag = 30.0286 A Vtrans = 0.300339 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:45:59 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.001155599 Vs 16:46:06 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.001171421 Vs 16:46:06 Integrated Voltage Average and RMS of Measurements X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.001163040 +- 0.000000289 Vs 16:46:06 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0000000 m, dx = 0.0100000 m, Nturns = 10.0 VT = 0.001163040 +- 0.000000289 Vs BL = 0.01163040 +- 0.00000289 Tm 16:46:10 Position moved: 0.000000 m 16:46:10 BL vs X Fit Coefficients: Cn sigCn n (Tm/m^n) ----- --------+-------- 0 0.01163 0.00000 Compute Fit Residuals: X0 BL sigBL BLfit BL-BLfit (m) (Tm) (Tm) (Tm) (Tm) ----------- --------+-------- -------- -------- 0.0000000 0.01163 0.00000 0.01163 0.00000 16:46:10 Magnet Strength Calculation Harmonic Number = 1 Number Of X0 Positions = 1 X0 BL sigBL (m) (Tm) (Tm) -------- --------+-------- 0.0000 0.01163 0.00000 SL = 0.0116304 +- 0.0000029 Tm^2/m^n 16:46:10 Magnet Strength Measurement Harmonic Number = 1 SL = 0.0116304 +- 0.0000029 Tm^2/m^n 16:46:13 Position moved: 0.000000 m 16:46:13 Magnet system MAIN current ramping to 15.0000 A... 16:46:25 Magnet Current for MAIN, Imag = 14.9977 A Vtrans = 0.150030 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:46:26 MAIN (V) Voltage Measurement, V = 0.588907 V 16:46:29 Temperature Measurement, Probe Number = 0, T = 26.041000 C 16:46:30 Temperature Measurement, Probe Number = 1, T = 24.109000 C 16:46:31 Temperature Measurement, Probe Number = 2, T = 27.326000 C 16:46:31 Temperature Measurement, Probe Number = 3, T = 28.737000 C 16:46:32 Temperature Measurement, Probe Number = 4, T = 28.707000 C 16:46:32 Temperature Measurement, Probe Number = 5, T = 24.501000 C 16:46:33 Sensor Voltage Measurement, Sensor Number = 0, V = 2.647975 V 16:46:33 Sensor Measurement, Sensor Number = 0, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 2.647975 (V, A, ohm, ...), Measured Value = 123.598125 (C, atm, Hz, ...) 16:46:34 Sensor Voltage Measurement, Sensor Number = 1, V = 1.154788 V 16:46:34 Sensor Measurement, Sensor Number = 1, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 1.154788 (V, A, ohm, ...), Measured Value = 11.609100 (C, atm, Hz, ...) 16:46:34 Magnet Test Parameter Measurements P0 = MAIN, Value = 15.00 P1 = MAIN (V), Value = 0.59 P2 = Yoke, Value = 26.04 P3 = Bottom Coil, Value = 24.11 P4 = Top Coil, Value = 27.33 P5 = Water In, Value = 28.74 P6 = Water Out, Value = 28.71 P7 = Ambient, Value = 24.50 P8 = Supply Pres(psi), Value = 123.60 P9 = Return Pres(psi), Value = 11.61 16:46:34 Magnetic Field Measurement Measuring the magnet strength... 16:46:34 Integrated Voltage Measurement... 16:46:34 Magnet Current for MAIN, Imag = 14.9975 A Vtrans = 0.150028 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:46:43 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.000578299 Vs 16:46:51 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.000593453 Vs 16:46:51 Magnet Current for MAIN, Imag = 14.9975 A Vtrans = 0.150028 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:46:59 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.000579378 Vs 16:47:07 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.000592180 Vs 16:47:07 Magnet Current for MAIN, Imag = 14.9976 A Vtrans = 0.150029 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:47:15 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.000579460 Vs 16:47:23 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.000593843 Vs 16:47:23 Magnet Current for MAIN, Imag = 14.9976 A Vtrans = 0.150029 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:47:31 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.000579176 Vs 16:47:39 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.000593109 Vs 16:47:39 Integrated Voltage Average and RMS of Measurements X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.000586112 +- 0.000000339 Vs 16:47:39 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0000000 m, dx = 0.0100000 m, Nturns = 10.0 VT = 0.000586112 +- 0.000000339 Vs BL = 0.00586112 +- 0.00000339 Tm 16:47:43 Position moved: 0.000000 m 16:47:43 BL vs X Fit Coefficients: Cn sigCn n (Tm/m^n) ----- --------+-------- 0 0.00586 0.00000 Compute Fit Residuals: X0 BL sigBL BLfit BL-BLfit (m) (Tm) (Tm) (Tm) (Tm) ----------- --------+-------- -------- -------- 0.0000000 0.00586 0.00000 0.00586 0.00000 16:47:43 Magnet Strength Calculation Harmonic Number = 1 Number Of X0 Positions = 1 X0 BL sigBL (m) (Tm) (Tm) -------- --------+-------- 0.0000 0.00586 0.00000 SL = 0.0058611 +- 0.0000034 Tm^2/m^n 16:47:43 Magnet Strength Measurement Harmonic Number = 1 SL = 0.0058611 +- 0.0000034 Tm^2/m^n Remeasure 16:47:43 Magnetic Field Measurement Measuring the magnet strength... 16:47:43 Integrated Voltage Measurement... 16:47:44 Magnet Current for MAIN, Imag = 14.9976 A Vtrans = 0.150030 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:47:52 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.000578478 Vs 16:48:00 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.000594360 Vs 16:48:01 Magnet Current for MAIN, Imag = 14.9976 A Vtrans = 0.150029 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:48:08 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.000577622 Vs 16:48:16 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.000593675 Vs 16:48:17 Magnet Current for MAIN, Imag = 14.9975 A Vtrans = 0.150028 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:48:24 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.000577667 Vs 16:48:32 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.000593972 Vs 16:48:33 Magnet Current for MAIN, Imag = 14.9976 A Vtrans = 0.150029 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:48:40 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.000576590 Vs 16:48:48 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.000594376 Vs 16:48:48 Integrated Voltage Average and RMS of Measurements X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.000585842 +- 0.000000353 Vs 16:48:48 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0000000 m, dx = 0.0100000 m, Nturns = 10.0 VT = 0.000585842 +- 0.000000353 Vs BL = 0.00585842 +- 0.00000353 Tm 16:48:51 Position moved: 0.000000 m 16:48:51 BL vs X Fit Coefficients: Cn sigCn n (Tm/m^n) ----- --------+-------- 0 0.00586 0.00000 Compute Fit Residuals: X0 BL sigBL BLfit BL-BLfit (m) (Tm) (Tm) (Tm) (Tm) ----------- --------+-------- -------- -------- 0.0000000 0.00586 0.00000 0.00586 0.00000 16:48:51 Magnet Strength Calculation Harmonic Number = 1 Number Of X0 Positions = 1 X0 BL sigBL (m) (Tm) (Tm) -------- --------+-------- 0.0000 0.00586 0.00000 SL = 0.0058584 +- 0.0000035 Tm^2/m^n 16:48:51 Magnet Strength Measurement Harmonic Number = 1 SL = 0.0058584 +- 0.0000035 Tm^2/m^n Remeasure 16:48:51 Magnetic Field Measurement Measuring the magnet strength... 16:48:51 Integrated Voltage Measurement... 16:48:52 Magnet Current for MAIN, Imag = 14.9976 A Vtrans = 0.150029 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:49:01 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.000578774 Vs 16:49:08 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.000594339 Vs 16:49:09 Magnet Current for MAIN, Imag = 14.9976 A Vtrans = 0.150029 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:49:17 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.000578574 Vs 16:49:24 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.000593103 Vs 16:49:25 Magnet Current for MAIN, Imag = 14.9975 A Vtrans = 0.150029 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:49:33 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.000578440 Vs 16:49:40 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.000593485 Vs 16:49:41 Magnet Current for MAIN, Imag = 14.9976 A Vtrans = 0.150030 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:49:49 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.000578456 Vs 16:49:56 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.000592715 Vs 16:49:56 Integrated Voltage Average and RMS of Measurements X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.000585986 +- 0.000000356 Vs 16:49:56 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0000000 m, dx = 0.0100000 m, Nturns = 10.0 VT = 0.000585986 +- 0.000000356 Vs BL = 0.00585986 +- 0.00000356 Tm 16:50:00 Position moved: 0.000000 m 16:50:00 BL vs X Fit Coefficients: Cn sigCn n (Tm/m^n) ----- --------+-------- 0 0.00586 0.00000 Compute Fit Residuals: X0 BL sigBL BLfit BL-BLfit (m) (Tm) (Tm) (Tm) (Tm) ----------- --------+-------- -------- -------- 0.0000000 0.00586 0.00000 0.00586 0.00000 16:50:00 Magnet Strength Calculation Harmonic Number = 1 Number Of X0 Positions = 1 X0 BL sigBL (m) (Tm) (Tm) -------- --------+-------- 0.0000 0.00586 0.00000 SL = 0.0058599 +- 0.0000036 Tm^2/m^n 16:50:00 Magnet Strength Measurement Harmonic Number = 1 SL = 0.0058599 +- 0.0000036 Tm^2/m^n Remeasure 16:50:00 Magnetic Field Measurement Measuring the magnet strength... 16:50:00 Integrated Voltage Measurement... 16:50:01 Magnet Current for MAIN, Imag = 14.9975 A Vtrans = 0.150029 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:50:09 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.000578447 Vs 16:50:17 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.000592870 Vs 16:50:18 Magnet Current for MAIN, Imag = 14.9976 A Vtrans = 0.150030 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:50:25 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.000578157 Vs 16:50:33 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.000593307 Vs 16:50:34 Magnet Current for MAIN, Imag = 14.9976 A Vtrans = 0.150030 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:50:41 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.000578239 Vs 16:50:49 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.000593485 Vs 16:50:50 Magnet Current for MAIN, Imag = 14.9976 A Vtrans = 0.150030 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:50:57 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.000578132 Vs 16:51:05 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.000594756 Vs 16:51:05 Integrated Voltage Average and RMS of Measurements X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.000585924 +- 0.000000309 Vs 16:51:05 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0000000 m, dx = 0.0100000 m, Nturns = 10.0 VT = 0.000585924 +- 0.000000309 Vs BL = 0.00585924 +- 0.00000309 Tm 16:51:09 Position moved: 0.000000 m 16:51:09 BL vs X Fit Coefficients: Cn sigCn n (Tm/m^n) ----- --------+-------- 0 0.00586 0.00000 Compute Fit Residuals: X0 BL sigBL BLfit BL-BLfit (m) (Tm) (Tm) (Tm) (Tm) ----------- --------+-------- -------- -------- 0.0000000 0.00586 0.00000 0.00586 0.00000 16:51:09 Magnet Strength Calculation Harmonic Number = 1 Number Of X0 Positions = 1 X0 BL sigBL (m) (Tm) (Tm) -------- --------+-------- 0.0000 0.00586 0.00000 SL = 0.0058592 +- 0.0000031 Tm^2/m^n 16:51:09 Magnet Strength Measurement Harmonic Number = 1 SL = 0.0058592 +- 0.0000031 Tm^2/m^n Remeasure 16:51:09 Magnetic Field Measurement Measuring the magnet strength... 16:51:09 Integrated Voltage Measurement... 16:51:10 Magnet Current for MAIN, Imag = 14.9976 A Vtrans = 0.150030 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:51:18 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.000577449 Vs 16:51:26 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.000594024 Vs 16:51:27 Magnet Current for MAIN, Imag = 14.9975 A Vtrans = 0.150029 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:51:34 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.000577194 Vs 16:51:42 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.000593763 Vs 16:51:43 Magnet Current for MAIN, Imag = 14.9976 A Vtrans = 0.150029 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:51:50 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.000578431 Vs 16:51:58 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.000593716 Vs 16:51:58 Magnet Current for MAIN, Imag = 14.9974 A Vtrans = 0.150027 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:52:06 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.000577801 Vs 16:52:14 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.000593969 Vs 16:52:14 Integrated Voltage Average and RMS of Measurements X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.000585793 +- 0.000000218 Vs 16:52:14 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0000000 m, dx = 0.0100000 m, Nturns = 10.0 VT = 0.000585793 +- 0.000000218 Vs BL = 0.00585793 +- 0.00000218 Tm 16:52:17 Position moved: 0.000000 m 16:52:17 BL vs X Fit Coefficients: Cn sigCn n (Tm/m^n) ----- --------+-------- 0 0.00586 0.00000 Compute Fit Residuals: X0 BL sigBL BLfit BL-BLfit (m) (Tm) (Tm) (Tm) (Tm) ----------- --------+-------- -------- -------- 0.0000000 0.00586 0.00000 0.00586 0.00000 16:52:17 Magnet Strength Calculation Harmonic Number = 1 Number Of X0 Positions = 1 X0 BL sigBL (m) (Tm) (Tm) -------- --------+-------- 0.0000 0.00586 0.00000 SL = 0.0058579 +- 0.0000022 Tm^2/m^n 16:52:17 Magnet Strength Measurement Harmonic Number = 1 SL = 0.0058579 +- 0.0000022 Tm^2/m^n 16:52:20 Position moved: 0.000000 m 16:52:20 Magnet system MAIN current ramping to 0.0000 A... 16:52:32 Magnet Current for MAIN, Imag = 0.0171 A Vtrans = 0.000225 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:52:33 MAIN (V) Voltage Measurement, V = 0.000671 V 16:52:37 Temperature Measurement, Probe Number = 0, T = 26.040000 C 16:52:37 Temperature Measurement, Probe Number = 1, T = 24.121000 C 16:52:38 Temperature Measurement, Probe Number = 2, T = 27.289000 C 16:52:38 Temperature Measurement, Probe Number = 3, T = 28.662000 C 16:52:39 Temperature Measurement, Probe Number = 4, T = 28.708000 C 16:52:40 Temperature Measurement, Probe Number = 5, T = 24.514000 C 16:52:40 Sensor Voltage Measurement, Sensor Number = 0, V = 2.658531 V 16:52:40 Sensor Measurement, Sensor Number = 0, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 2.658531 (V, A, ohm, ...), Measured Value = 124.389840 (C, atm, Hz, ...) 16:52:41 Sensor Voltage Measurement, Sensor Number = 1, V = 1.156587 V 16:52:41 Sensor Measurement, Sensor Number = 1, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 1.156587 (V, A, ohm, ...), Measured Value = 11.744047 (C, atm, Hz, ...) 16:52:41 Magnet Test Parameter Measurements P0 = MAIN, Value = 0.02 P1 = MAIN (V), Value = 0.00 P2 = Yoke, Value = 26.04 P3 = Bottom Coil, Value = 24.12 P4 = Top Coil, Value = 27.29 P5 = Water In, Value = 28.66 P6 = Water Out, Value = 28.71 P7 = Ambient, Value = 24.51 P8 = Supply Pres(psi), Value = 124.39 P9 = Return Pres(psi), Value = 11.74 16:52:41 Magnetic Field Measurement Measuring the magnet strength... 16:52:41 Integrated Voltage Measurement... 16:52:41 Magnet Current for MAIN, Imag = 0.0173 A Vtrans = 0.000226 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:52:50 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.000001962 Vs 16:52:58 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.000017345 Vs 16:52:58 Magnet Current for MAIN, Imag = 0.0170 A Vtrans = 0.000224 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:53:06 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.000000901 Vs 16:53:14 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.000016746 Vs 16:53:14 Magnet Current for MAIN, Imag = 0.0171 A Vtrans = 0.000224 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:53:22 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.000000878 Vs 16:53:30 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.000017638 Vs 16:53:30 Magnet Current for MAIN, Imag = 0.0170 A Vtrans = 0.000224 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:53:38 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.000001440 Vs 16:53:46 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0100000 m VT = -0.000017035 Vs 16:53:46 Integrated Voltage Average and RMS of Measurements X0 = 0.0000000 m, Delta X = 0.0100000 m VT = 0.000009243 +- 0.000000293 Vs 16:53:46 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0000000 m, dx = 0.0100000 m, Nturns = 10.0 VT = 0.000009243 +- 0.000000293 Vs BL = 0.00009243 +- 0.00000293 Tm 16:53:49 Position moved: 0.000000 m 16:53:49 BL vs X Fit Coefficients: Cn sigCn n (Tm/m^n) ----- --------+-------- 0 0.00009 0.00000 Compute Fit Residuals: X0 BL sigBL BLfit BL-BLfit (m) (Tm) (Tm) (Tm) (Tm) ----------- --------+-------- -------- -------- 0.0000000 0.00009 0.00000 0.00009 0.00000 16:53:49 Magnet Strength Calculation Harmonic Number = 1 Number Of X0 Positions = 1 X0 BL sigBL (m) (Tm) (Tm) -------- --------+-------- 0.0000 0.00009 0.00000 SL = 0.0000924 +- 0.0000029 Tm^2/m^n 16:53:49 Magnet Strength Measurement Harmonic Number = 1 SL = 0.0000924 +- 0.0000029 Tm^2/m^n 16:53:52 Position moved: 0.000000 m 16:53:52 Magnet system MAIN current ramping to 155.0000 A... 16:54:14 Magnet Current for MAIN, Imag = 154.9747 A Vtrans = 1.549800 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:54:14 MAIN (V) Voltage Measurement, V = 6.111358 V 16:54:18 Temperature Measurement, Probe Number = 0, T = 26.034000 C 16:54:19 Temperature Measurement, Probe Number = 1, T = 24.109000 C 16:54:19 Temperature Measurement, Probe Number = 2, T = 27.384000 C 16:54:20 Temperature Measurement, Probe Number = 3, T = 28.709000 C 16:54:20 Temperature Measurement, Probe Number = 4, T = 28.718000 C 16:54:21 Temperature Measurement, Probe Number = 5, T = 24.468000 C 16:54:22 Sensor Voltage Measurement, Sensor Number = 0, V = 2.664537 V 16:54:22 Sensor Measurement, Sensor Number = 0, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 2.664537 (V, A, ohm, ...), Measured Value = 124.840298 (C, atm, Hz, ...) 16:54:22 Sensor Voltage Measurement, Sensor Number = 1, V = 1.156378 V 16:54:22 Sensor Measurement, Sensor Number = 1, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 1.156378 (V, A, ohm, ...), Measured Value = 11.728313 (C, atm, Hz, ...) 16:54:22 Magnet Test Parameter Measurements P0 = MAIN, Value = 154.97 P1 = MAIN (V), Value = 6.11 P2 = Yoke, Value = 26.03 P3 = Bottom Coil, Value = 24.11 P4 = Top Coil, Value = 27.38 P5 = Water In, Value = 28.71 P6 = Water Out, Value = 28.72 P7 = Ambient, Value = 24.47 P8 = Supply Pres(psi), Value = 124.84 P9 = Return Pres(psi), Value = 11.73 16:54:22 Integrated Voltage Measurement... 16:54:23 Magnet Current for MAIN, Imag = 154.9750 A Vtrans = 1.549803 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:54:32 Integrated Voltage Measurement X0 = -0.0350000 m, Delta X = 0.0050000 m VT = 0.002648434 Vs 16:54:39 Integrated Voltage Measurement X0 = -0.0350000 m, Delta X = -0.0050000 m VT = -0.002667068 Vs 16:54:40 Magnet Current for MAIN, Imag = 154.9750 A Vtrans = 1.549803 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:54:48 Integrated Voltage Measurement X0 = -0.0350000 m, Delta X = 0.0050000 m VT = 0.002660271 Vs 16:54:55 Integrated Voltage Measurement X0 = -0.0350000 m, Delta X = -0.0050000 m VT = -0.002670006 Vs 16:54:56 Magnet Current for MAIN, Imag = 154.9752 A Vtrans = 1.549805 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:55:03 Integrated Voltage Measurement X0 = -0.0350000 m, Delta X = 0.0050000 m VT = 0.002652253 Vs 16:55:11 Integrated Voltage Measurement X0 = -0.0350000 m, Delta X = -0.0050000 m VT = -0.002665924 Vs 16:55:12 Magnet Current for MAIN, Imag = 154.9746 A Vtrans = 1.549799 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:55:19 Integrated Voltage Measurement X0 = -0.0350000 m, Delta X = 0.0050000 m VT = 0.002649227 Vs 16:55:27 Integrated Voltage Measurement X0 = -0.0350000 m, Delta X = -0.0050000 m VT = -0.002660508 Vs 16:55:27 Integrated Voltage Average and RMS of Measurements X0 = -0.0350000 m, Delta X = 0.0050000 m VT = 0.002659211 +- 0.000003747 Vs 16:55:27 Integrated Field Strength, Average and RMS of Measurements X0 = -0.0350000 m, dx = 0.0050000 m, Nturns = 10.0 VT = 0.002659211 +- 0.000003747 Vs BL = 0.05318423 +- 0.00007493 Tm Remeasure 16:55:27 Integrated Voltage Measurement... 16:55:27 Magnet Current for MAIN, Imag = 154.9747 A Vtrans = 1.549800 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:55:35 Integrated Voltage Measurement X0 = -0.0350000 m, Delta X = 0.0050000 m VT = 0.002651183 Vs 16:55:43 Integrated Voltage Measurement X0 = -0.0350000 m, Delta X = -0.0050000 m VT = -0.002665277 Vs 16:55:43 Magnet Current for MAIN, Imag = 154.9752 A Vtrans = 1.549805 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:55:51 Integrated Voltage Measurement X0 = -0.0350000 m, Delta X = 0.0050000 m VT = 0.002653408 Vs 16:55:59 Integrated Voltage Measurement X0 = -0.0350000 m, Delta X = -0.0050000 m VT = -0.002666252 Vs 16:55:59 Magnet Current for MAIN, Imag = 154.9752 A Vtrans = 1.549805 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:56:07 Integrated Voltage Measurement X0 = -0.0350000 m, Delta X = 0.0050000 m VT = 0.002646934 Vs 16:56:15 Integrated Voltage Measurement X0 = -0.0350000 m, Delta X = -0.0050000 m VT = -0.002673661 Vs 16:56:15 Magnet Current for MAIN, Imag = 154.9746 A Vtrans = 1.549799 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:56:23 Integrated Voltage Measurement X0 = -0.0350000 m, Delta X = 0.0050000 m VT = 0.002654602 Vs 16:56:31 Integrated Voltage Measurement X0 = -0.0350000 m, Delta X = -0.0050000 m VT = -0.002665627 Vs 16:56:31 Integrated Voltage Average and RMS of Measurements X0 = -0.0350000 m, Delta X = 0.0050000 m VT = 0.002659618 +- 0.000000818 Vs 16:56:31 Integrated Field Strength, Average and RMS of Measurements X0 = -0.0350000 m, dx = 0.0050000 m, Nturns = 10.0 VT = 0.002659618 +- 0.000000818 Vs BL = 0.05319236 +- 0.00001637 Tm Remeasure 16:56:31 Integrated Voltage Measurement... 16:56:31 Magnet Current for MAIN, Imag = 154.9748 A Vtrans = 1.549802 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:56:39 Integrated Voltage Measurement X0 = -0.0350000 m, Delta X = 0.0050000 m VT = 0.002659446 Vs 16:56:47 Integrated Voltage Measurement X0 = -0.0350000 m, Delta X = -0.0050000 m VT = -0.002672298 Vs 16:56:47 Magnet Current for MAIN, Imag = 154.9748 A Vtrans = 1.549802 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:56:55 Integrated Voltage Measurement X0 = -0.0350000 m, Delta X = 0.0050000 m VT = 0.002657118 Vs 16:57:03 Integrated Voltage Measurement X0 = -0.0350000 m, Delta X = -0.0050000 m VT = -0.002671193 Vs 16:57:03 Magnet Current for MAIN, Imag = 154.9745 A Vtrans = 1.549798 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:57:11 Integrated Voltage Measurement X0 = -0.0350000 m, Delta X = 0.0050000 m VT = 0.002650288 Vs 16:57:19 Integrated Voltage Measurement X0 = -0.0350000 m, Delta X = -0.0050000 m VT = -0.002668312 Vs 16:57:19 Magnet Current for MAIN, Imag = 154.9745 A Vtrans = 1.549798 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:57:27 Integrated Voltage Measurement X0 = -0.0350000 m, Delta X = 0.0050000 m VT = 0.002659086 Vs 16:57:35 Integrated Voltage Measurement X0 = -0.0350000 m, Delta X = -0.0050000 m VT = -0.002677579 Vs 16:57:35 Integrated Voltage Average and RMS of Measurements X0 = -0.0350000 m, Delta X = 0.0050000 m VT = 0.002664415 +- 0.000003305 Vs 16:57:35 Integrated Field Strength, Average and RMS of Measurements X0 = -0.0350000 m, dx = 0.0050000 m, Nturns = 10.0 VT = 0.002664415 +- 0.000003305 Vs BL = 0.05328830 +- 0.00006611 Tm Remeasure 16:57:35 Integrated Voltage Measurement... 16:57:35 Magnet Current for MAIN, Imag = 154.9748 A Vtrans = 1.549802 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:57:43 Integrated Voltage Measurement X0 = -0.0350000 m, Delta X = 0.0050000 m VT = 0.002646937 Vs 16:57:51 Integrated Voltage Measurement X0 = -0.0350000 m, Delta X = -0.0050000 m VT = -0.002676627 Vs 16:57:51 Magnet Current for MAIN, Imag = 154.9746 A Vtrans = 1.549799 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:57:59 Integrated Voltage Measurement X0 = -0.0350000 m, Delta X = 0.0050000 m VT = 0.002660127 Vs 16:58:06 Integrated Voltage Measurement X0 = -0.0350000 m, Delta X = -0.0050000 m VT = -0.002670954 Vs 16:58:07 Magnet Current for MAIN, Imag = 154.9750 A Vtrans = 1.549803 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:58:15 Integrated Voltage Measurement X0 = -0.0350000 m, Delta X = 0.0050000 m VT = 0.002651514 Vs 16:58:22 Integrated Voltage Measurement X0 = -0.0350000 m, Delta X = -0.0050000 m VT = -0.002660046 Vs 16:58:23 Magnet Current for MAIN, Imag = 154.9751 A Vtrans = 1.549804 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:58:31 Integrated Voltage Measurement X0 = -0.0350000 m, Delta X = 0.0050000 m VT = 0.002646251 Vs 16:58:38 Integrated Voltage Measurement X0 = -0.0350000 m, Delta X = -0.0050000 m VT = -0.002667340 Vs 16:58:38 Integrated Voltage Average and RMS of Measurements X0 = -0.0350000 m, Delta X = 0.0050000 m VT = 0.002659975 +- 0.000003935 Vs 16:58:38 Integrated Field Strength, Average and RMS of Measurements X0 = -0.0350000 m, dx = 0.0050000 m, Nturns = 10.0 VT = 0.002659975 +- 0.000003935 Vs BL = 0.05319949 +- 0.00007871 Tm Remeasure 16:58:38 Integrated Voltage Measurement... 16:58:39 Magnet Current for MAIN, Imag = 154.9752 A Vtrans = 1.549805 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:58:47 Integrated Voltage Measurement X0 = -0.0350000 m, Delta X = 0.0050000 m VT = 0.002664428 Vs 16:58:55 Integrated Voltage Measurement X0 = -0.0350000 m, Delta X = -0.0050000 m VT = -0.002664987 Vs 16:58:55 Magnet Current for MAIN, Imag = 154.9745 A Vtrans = 1.549798 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:59:03 Integrated Voltage Measurement X0 = -0.0350000 m, Delta X = 0.0050000 m VT = 0.002658864 Vs 16:59:10 Integrated Voltage Measurement X0 = -0.0350000 m, Delta X = -0.0050000 m VT = -0.002669538 Vs 16:59:11 Magnet Current for MAIN, Imag = 154.9748 A Vtrans = 1.549802 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:59:19 Integrated Voltage Measurement X0 = -0.0350000 m, Delta X = 0.0050000 m VT = 0.002647778 Vs 16:59:26 Integrated Voltage Measurement X0 = -0.0350000 m, Delta X = -0.0050000 m VT = -0.002662158 Vs 16:59:27 Magnet Current for MAIN, Imag = 154.9752 A Vtrans = 1.549805 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:59:35 Integrated Voltage Measurement X0 = -0.0350000 m, Delta X = 0.0050000 m VT = 0.002658877 Vs 16:59:42 Integrated Voltage Measurement X0 = -0.0350000 m, Delta X = -0.0050000 m VT = -0.002670319 Vs 16:59:42 Integrated Voltage Average and RMS of Measurements X0 = -0.0350000 m, Delta X = 0.0050000 m VT = 0.002662119 +- 0.000004133 Vs 16:59:42 Integrated Field Strength, Average and RMS of Measurements X0 = -0.0350000 m, dx = 0.0050000 m, Nturns = 10.0 VT = 0.002662119 +- 0.000004133 Vs BL = 0.05324237 +- 0.00008265 Tm 16:59:42 Integrated Voltage Measurement... 16:59:43 Magnet Current for MAIN, Imag = 154.9752 A Vtrans = 1.549805 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 16:59:52 Integrated Voltage Measurement X0 = -0.0300000 m, Delta X = 0.0050000 m VT = 0.002816960 Vs 16:59:59 Integrated Voltage Measurement X0 = -0.0300000 m, Delta X = -0.0050000 m VT = -0.002838845 Vs 17:00:00 Magnet Current for MAIN, Imag = 154.9750 A Vtrans = 1.549803 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:00:08 Integrated Voltage Measurement X0 = -0.0300000 m, Delta X = 0.0050000 m VT = 0.002821347 Vs 17:00:15 Integrated Voltage Measurement X0 = -0.0300000 m, Delta X = -0.0050000 m VT = -0.002838173 Vs 17:00:16 Magnet Current for MAIN, Imag = 154.9751 A Vtrans = 1.549804 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:00:24 Integrated Voltage Measurement X0 = -0.0300000 m, Delta X = 0.0050000 m VT = 0.002818128 Vs 17:00:31 Integrated Voltage Measurement X0 = -0.0300000 m, Delta X = -0.0050000 m VT = -0.002842371 Vs 17:00:32 Magnet Current for MAIN, Imag = 154.9750 A Vtrans = 1.549803 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:00:39 Integrated Voltage Measurement X0 = -0.0300000 m, Delta X = 0.0050000 m VT = 0.002816523 Vs 17:00:47 Integrated Voltage Measurement X0 = -0.0300000 m, Delta X = -0.0050000 m VT = -0.002824145 Vs 17:00:47 Integrated Voltage Average and RMS of Measurements X0 = -0.0300000 m, Delta X = 0.0050000 m VT = 0.002827062 +- 0.000003982 Vs 17:00:47 Integrated Field Strength, Average and RMS of Measurements X0 = -0.0300000 m, dx = 0.0050000 m, Nturns = 10.0 VT = 0.002827062 +- 0.000003982 Vs BL = 0.05654123 +- 0.00007963 Tm Remeasure 17:00:47 Integrated Voltage Measurement... 17:00:48 Magnet Current for MAIN, Imag = 154.9747 A Vtrans = 1.549800 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:00:55 Integrated Voltage Measurement X0 = -0.0300000 m, Delta X = 0.0050000 m VT = 0.002823631 Vs 17:01:03 Integrated Voltage Measurement X0 = -0.0300000 m, Delta X = -0.0050000 m VT = -0.002826809 Vs 17:01:04 Magnet Current for MAIN, Imag = 154.9750 A Vtrans = 1.549803 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:01:11 Integrated Voltage Measurement X0 = -0.0300000 m, Delta X = 0.0050000 m VT = 0.002819479 Vs 17:01:19 Integrated Voltage Measurement X0 = -0.0300000 m, Delta X = -0.0050000 m VT = -0.002834741 Vs 17:01:20 Magnet Current for MAIN, Imag = 154.9747 A Vtrans = 1.549800 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:01:27 Integrated Voltage Measurement X0 = -0.0300000 m, Delta X = 0.0050000 m VT = 0.002815301 Vs 17:01:35 Integrated Voltage Measurement X0 = -0.0300000 m, Delta X = -0.0050000 m VT = -0.002830784 Vs 17:01:35 Magnet Current for MAIN, Imag = 154.9743 A Vtrans = 1.549797 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:01:43 Integrated Voltage Measurement X0 = -0.0300000 m, Delta X = 0.0050000 m VT = 0.002814206 Vs 17:01:51 Integrated Voltage Measurement X0 = -0.0300000 m, Delta X = -0.0050000 m VT = -0.002838158 Vs 17:01:51 Integrated Voltage Average and RMS of Measurements X0 = -0.0300000 m, Delta X = 0.0050000 m VT = 0.002825388 +- 0.000001510 Vs 17:01:51 Integrated Field Strength, Average and RMS of Measurements X0 = -0.0300000 m, dx = 0.0050000 m, Nturns = 10.0 VT = 0.002825388 +- 0.000001510 Vs BL = 0.05650777 +- 0.00003021 Tm Remeasure 17:01:51 Integrated Voltage Measurement... 17:01:52 Magnet Current for MAIN, Imag = 154.9745 A Vtrans = 1.549798 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:01:59 Integrated Voltage Measurement X0 = -0.0300000 m, Delta X = 0.0050000 m VT = 0.002816921 Vs 17:02:07 Integrated Voltage Measurement X0 = -0.0300000 m, Delta X = -0.0050000 m VT = -0.002828548 Vs 17:02:07 Magnet Current for MAIN, Imag = 154.9748 A Vtrans = 1.549802 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:02:15 Integrated Voltage Measurement X0 = -0.0300000 m, Delta X = 0.0050000 m VT = 0.002821462 Vs 17:02:23 Integrated Voltage Measurement X0 = -0.0300000 m, Delta X = -0.0050000 m VT = -0.002823682 Vs 17:02:23 Magnet Current for MAIN, Imag = 154.9745 A Vtrans = 1.549798 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:02:31 Integrated Voltage Measurement X0 = -0.0300000 m, Delta X = 0.0050000 m VT = 0.002814681 Vs 17:02:38 Integrated Voltage Measurement X0 = -0.0300000 m, Delta X = -0.0050000 m VT = -0.002836986 Vs 17:02:39 Magnet Current for MAIN, Imag = 154.9752 A Vtrans = 1.549805 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:02:47 Integrated Voltage Measurement X0 = -0.0300000 m, Delta X = 0.0050000 m VT = 0.002818202 Vs 17:02:54 Integrated Voltage Measurement X0 = -0.0300000 m, Delta X = -0.0050000 m VT = -0.002826902 Vs 17:02:54 Integrated Voltage Average and RMS of Measurements X0 = -0.0300000 m, Delta X = 0.0050000 m VT = 0.002823423 +- 0.000001393 Vs 17:02:54 Integrated Field Strength, Average and RMS of Measurements X0 = -0.0300000 m, dx = 0.0050000 m, Nturns = 10.0 VT = 0.002823423 +- 0.000001393 Vs BL = 0.05646846 +- 0.00002787 Tm Remeasure 17:02:54 Integrated Voltage Measurement... 17:02:55 Magnet Current for MAIN, Imag = 154.9751 A Vtrans = 1.549804 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:03:03 Integrated Voltage Measurement X0 = -0.0300000 m, Delta X = 0.0050000 m VT = 0.002815355 Vs 17:03:10 Integrated Voltage Measurement X0 = -0.0300000 m, Delta X = -0.0050000 m VT = -0.002834386 Vs 17:03:11 Magnet Current for MAIN, Imag = 154.9750 A Vtrans = 1.549803 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:03:18 Integrated Voltage Measurement X0 = -0.0300000 m, Delta X = 0.0050000 m VT = 0.002813704 Vs 17:03:26 Integrated Voltage Measurement X0 = -0.0300000 m, Delta X = -0.0050000 m VT = -0.002834025 Vs 17:03:27 Magnet Current for MAIN, Imag = 154.9750 A Vtrans = 1.549803 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:03:34 Integrated Voltage Measurement X0 = -0.0300000 m, Delta X = 0.0050000 m VT = 0.002821276 Vs 17:03:42 Integrated Voltage Measurement X0 = -0.0300000 m, Delta X = -0.0050000 m VT = -0.002832550 Vs 17:03:43 Magnet Current for MAIN, Imag = 154.9753 A Vtrans = 1.549807 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:03:50 Integrated Voltage Measurement X0 = -0.0300000 m, Delta X = 0.0050000 m VT = 0.002817165 Vs 17:03:58 Integrated Voltage Measurement X0 = -0.0300000 m, Delta X = -0.0050000 m VT = -0.002825078 Vs 17:03:58 Integrated Voltage Average and RMS of Measurements X0 = -0.0300000 m, Delta X = 0.0050000 m VT = 0.002824192 +- 0.000002086 Vs 17:03:58 Integrated Field Strength, Average and RMS of Measurements X0 = -0.0300000 m, dx = 0.0050000 m, Nturns = 10.0 VT = 0.002824192 +- 0.000002086 Vs BL = 0.05648385 +- 0.00004171 Tm Remeasure 17:03:58 Integrated Voltage Measurement... 17:03:59 Magnet Current for MAIN, Imag = 154.9748 A Vtrans = 1.549802 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:04:06 Integrated Voltage Measurement X0 = -0.0300000 m, Delta X = 0.0050000 m VT = 0.002815629 Vs 17:04:14 Integrated Voltage Measurement X0 = -0.0300000 m, Delta X = -0.0050000 m VT = -0.002841929 Vs 17:04:14 Magnet Current for MAIN, Imag = 154.9746 A Vtrans = 1.549799 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:04:22 Integrated Voltage Measurement X0 = -0.0300000 m, Delta X = 0.0050000 m VT = 0.002818877 Vs 17:04:30 Integrated Voltage Measurement X0 = -0.0300000 m, Delta X = -0.0050000 m VT = -0.002834575 Vs 17:04:30 Magnet Current for MAIN, Imag = 154.9750 A Vtrans = 1.549803 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:04:38 Integrated Voltage Measurement X0 = -0.0300000 m, Delta X = 0.0050000 m VT = 0.002817522 Vs 17:04:46 Integrated Voltage Measurement X0 = -0.0300000 m, Delta X = -0.0050000 m VT = -0.002838172 Vs 17:04:46 Magnet Current for MAIN, Imag = 154.9747 A Vtrans = 1.549800 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:04:54 Integrated Voltage Measurement X0 = -0.0300000 m, Delta X = 0.0050000 m VT = 0.002815277 Vs 17:05:02 Integrated Voltage Measurement X0 = -0.0300000 m, Delta X = -0.0050000 m VT = -0.002833257 Vs 17:05:02 Integrated Voltage Average and RMS of Measurements X0 = -0.0300000 m, Delta X = 0.0050000 m VT = 0.002826905 +- 0.000001688 Vs 17:05:02 Integrated Field Strength, Average and RMS of Measurements X0 = -0.0300000 m, dx = 0.0050000 m, Nturns = 10.0 VT = 0.002826905 +- 0.000001688 Vs BL = 0.05653809 +- 0.00003375 Tm 17:05:02 Integrated Voltage Measurement... 17:05:02 Magnet Current for MAIN, Imag = 154.9746 A Vtrans = 1.549799 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:05:11 Integrated Voltage Measurement X0 = -0.0250000 m, Delta X = 0.0050000 m VT = 0.002903616 Vs 17:05:19 Integrated Voltage Measurement X0 = -0.0250000 m, Delta X = -0.0050000 m VT = -0.002912443 Vs 17:05:19 Magnet Current for MAIN, Imag = 154.9748 A Vtrans = 1.549802 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:05:27 Integrated Voltage Measurement X0 = -0.0250000 m, Delta X = 0.0050000 m VT = 0.002899467 Vs 17:05:35 Integrated Voltage Measurement X0 = -0.0250000 m, Delta X = -0.0050000 m VT = -0.002920559 Vs 17:05:35 Magnet Current for MAIN, Imag = 154.9751 A Vtrans = 1.549804 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:05:43 Integrated Voltage Measurement X0 = -0.0250000 m, Delta X = 0.0050000 m VT = 0.002900152 Vs 17:05:51 Integrated Voltage Measurement X0 = -0.0250000 m, Delta X = -0.0050000 m VT = -0.002916399 Vs 17:05:51 Magnet Current for MAIN, Imag = 154.9751 A Vtrans = 1.549804 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:05:59 Integrated Voltage Measurement X0 = -0.0250000 m, Delta X = 0.0050000 m VT = 0.002896943 Vs 17:06:06 Integrated Voltage Measurement X0 = -0.0250000 m, Delta X = -0.0050000 m VT = -0.002915530 Vs 17:06:06 Integrated Voltage Average and RMS of Measurements X0 = -0.0250000 m, Delta X = 0.0050000 m VT = 0.002908139 +- 0.000001338 Vs 17:06:06 Integrated Field Strength, Average and RMS of Measurements X0 = -0.0250000 m, dx = 0.0050000 m, Nturns = 10.0 VT = 0.002908139 +- 0.000001338 Vs BL = 0.05816277 +- 0.00002676 Tm Remeasure 17:06:06 Integrated Voltage Measurement... 17:06:07 Magnet Current for MAIN, Imag = 154.9750 A Vtrans = 1.549803 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:06:15 Integrated Voltage Measurement X0 = -0.0250000 m, Delta X = 0.0050000 m VT = 0.002903004 Vs 17:06:22 Integrated Voltage Measurement X0 = -0.0250000 m, Delta X = -0.0050000 m VT = -0.002923371 Vs 17:06:23 Magnet Current for MAIN, Imag = 154.9751 A Vtrans = 1.549804 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:06:31 Integrated Voltage Measurement X0 = -0.0250000 m, Delta X = 0.0050000 m VT = 0.002898723 Vs 17:06:38 Integrated Voltage Measurement X0 = -0.0250000 m, Delta X = -0.0050000 m VT = -0.002914481 Vs 17:06:39 Magnet Current for MAIN, Imag = 154.9751 A Vtrans = 1.549804 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:06:47 Integrated Voltage Measurement X0 = -0.0250000 m, Delta X = 0.0050000 m VT = 0.002898124 Vs 17:06:54 Integrated Voltage Measurement X0 = -0.0250000 m, Delta X = -0.0050000 m VT = -0.002911788 Vs 17:06:55 Magnet Current for MAIN, Imag = 154.9752 A Vtrans = 1.549805 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:07:03 Integrated Voltage Measurement X0 = -0.0250000 m, Delta X = 0.0050000 m VT = 0.002896242 Vs 17:07:10 Integrated Voltage Measurement X0 = -0.0250000 m, Delta X = -0.0050000 m VT = -0.002919869 Vs 17:07:10 Integrated Voltage Average and RMS of Measurements X0 = -0.0250000 m, Delta X = 0.0050000 m VT = 0.002908200 +- 0.000003081 Vs 17:07:10 Integrated Field Strength, Average and RMS of Measurements X0 = -0.0250000 m, dx = 0.0050000 m, Nturns = 10.0 VT = 0.002908200 +- 0.000003081 Vs BL = 0.05816401 +- 0.00006162 Tm Remeasure 17:07:10 Integrated Voltage Measurement... 17:07:11 Magnet Current for MAIN, Imag = 154.9752 A Vtrans = 1.549805 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:07:19 Integrated Voltage Measurement X0 = -0.0250000 m, Delta X = 0.0050000 m VT = 0.002900020 Vs 17:07:26 Integrated Voltage Measurement X0 = -0.0250000 m, Delta X = -0.0050000 m VT = -0.002920340 Vs 17:07:27 Magnet Current for MAIN, Imag = 154.9748 A Vtrans = 1.549802 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:07:35 Integrated Voltage Measurement X0 = -0.0250000 m, Delta X = 0.0050000 m VT = 0.002903967 Vs 17:07:42 Integrated Voltage Measurement X0 = -0.0250000 m, Delta X = -0.0050000 m VT = -0.002922196 Vs 17:07:43 Magnet Current for MAIN, Imag = 154.9748 A Vtrans = 1.549802 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:07:51 Integrated Voltage Measurement X0 = -0.0250000 m, Delta X = 0.0050000 m VT = 0.002903331 Vs 17:07:58 Integrated Voltage Measurement X0 = -0.0250000 m, Delta X = -0.0050000 m VT = -0.002917018 Vs 17:07:59 Magnet Current for MAIN, Imag = 154.9753 A Vtrans = 1.549807 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:08:07 Integrated Voltage Measurement X0 = -0.0250000 m, Delta X = 0.0050000 m VT = 0.002898884 Vs 17:08:14 Integrated Voltage Measurement X0 = -0.0250000 m, Delta X = -0.0050000 m VT = -0.002914542 Vs 17:08:14 Integrated Voltage Average and RMS of Measurements X0 = -0.0250000 m, Delta X = 0.0050000 m VT = 0.002910037 +- 0.000002256 Vs 17:08:14 Integrated Field Strength, Average and RMS of Measurements X0 = -0.0250000 m, dx = 0.0050000 m, Nturns = 10.0 VT = 0.002910037 +- 0.000002256 Vs BL = 0.05820074 +- 0.00004512 Tm Remeasure 17:08:14 Integrated Voltage Measurement... 17:08:15 Magnet Current for MAIN, Imag = 154.9750 A Vtrans = 1.549803 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:08:22 Integrated Voltage Measurement X0 = -0.0250000 m, Delta X = 0.0050000 m VT = 0.002901895 Vs 17:08:30 Integrated Voltage Measurement X0 = -0.0250000 m, Delta X = -0.0050000 m VT = -0.002920885 Vs 17:08:31 Magnet Current for MAIN, Imag = 154.9752 A Vtrans = 1.549805 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:08:38 Integrated Voltage Measurement X0 = -0.0250000 m, Delta X = 0.0050000 m VT = 0.002898986 Vs 17:08:46 Integrated Voltage Measurement X0 = -0.0250000 m, Delta X = -0.0050000 m VT = -0.002910819 Vs 17:08:46 Magnet Current for MAIN, Imag = 154.9750 A Vtrans = 1.549803 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:08:54 Integrated Voltage Measurement X0 = -0.0250000 m, Delta X = 0.0050000 m VT = 0.002899243 Vs 17:09:02 Integrated Voltage Measurement X0 = -0.0250000 m, Delta X = -0.0050000 m VT = -0.002909392 Vs 17:09:02 Magnet Current for MAIN, Imag = 154.9750 A Vtrans = 1.549803 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:09:10 Integrated Voltage Measurement X0 = -0.0250000 m, Delta X = 0.0050000 m VT = 0.002904206 Vs 17:09:18 Integrated Voltage Measurement X0 = -0.0250000 m, Delta X = -0.0050000 m VT = -0.002917383 Vs 17:09:18 Integrated Voltage Average and RMS of Measurements X0 = -0.0250000 m, Delta X = 0.0050000 m VT = 0.002907851 +- 0.000003254 Vs 17:09:18 Integrated Field Strength, Average and RMS of Measurements X0 = -0.0250000 m, dx = 0.0050000 m, Nturns = 10.0 VT = 0.002907851 +- 0.000003254 Vs BL = 0.05815702 +- 0.00006509 Tm Remeasure 17:09:18 Integrated Voltage Measurement... 17:09:18 Magnet Current for MAIN, Imag = 154.9755 A Vtrans = 1.549808 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:09:26 Integrated Voltage Measurement X0 = -0.0250000 m, Delta X = 0.0050000 m VT = 0.002903127 Vs 17:09:34 Integrated Voltage Measurement X0 = -0.0250000 m, Delta X = -0.0050000 m VT = -0.002913592 Vs 17:09:34 Magnet Current for MAIN, Imag = 154.9750 A Vtrans = 1.549803 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:09:42 Integrated Voltage Measurement X0 = -0.0250000 m, Delta X = 0.0050000 m VT = 0.002900781 Vs 17:09:49 Integrated Voltage Measurement X0 = -0.0250000 m, Delta X = -0.0050000 m VT = -0.002908093 Vs 17:09:50 Magnet Current for MAIN, Imag = 154.9748 A Vtrans = 1.549802 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:09:58 Integrated Voltage Measurement X0 = -0.0250000 m, Delta X = 0.0050000 m VT = 0.002903689 Vs 17:10:05 Integrated Voltage Measurement X0 = -0.0250000 m, Delta X = -0.0050000 m VT = -0.002919936 Vs 17:10:06 Magnet Current for MAIN, Imag = 154.9750 A Vtrans = 1.549803 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:10:14 Integrated Voltage Measurement X0 = -0.0250000 m, Delta X = 0.0050000 m VT = 0.002900097 Vs 17:10:21 Integrated Voltage Measurement X0 = -0.0250000 m, Delta X = -0.0050000 m VT = -0.002909002 Vs 17:10:21 Integrated Voltage Average and RMS of Measurements X0 = -0.0250000 m, Delta X = 0.0050000 m VT = 0.002907290 +- 0.000003051 Vs 17:10:21 Integrated Field Strength, Average and RMS of Measurements X0 = -0.0250000 m, dx = 0.0050000 m, Nturns = 10.0 VT = 0.002907290 +- 0.000003051 Vs BL = 0.05814579 +- 0.00006103 Tm 17:10:21 Integrated Voltage Measurement... 17:10:22 Magnet Current for MAIN, Imag = 154.9752 A Vtrans = 1.549805 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:10:30 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = 0.0050000 m VT = 0.002938252 Vs 17:10:38 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = -0.0050000 m VT = -0.002955901 Vs 17:10:39 Magnet Current for MAIN, Imag = 154.9752 A Vtrans = 1.549805 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:10:46 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = 0.0050000 m VT = 0.002940114 Vs 17:10:54 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = -0.0050000 m VT = -0.002954005 Vs 17:10:55 Magnet Current for MAIN, Imag = 154.9751 A Vtrans = 1.549804 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:11:02 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = 0.0050000 m VT = 0.002939791 Vs 17:11:10 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = -0.0050000 m VT = -0.002952427 Vs 17:11:11 Magnet Current for MAIN, Imag = 154.9752 A Vtrans = 1.549805 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:11:18 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = 0.0050000 m VT = 0.002938532 Vs 17:11:26 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = -0.0050000 m VT = -0.002949950 Vs 17:11:26 Integrated Voltage Average and RMS of Measurements X0 = -0.0200000 m, Delta X = 0.0050000 m VT = 0.002946121 +- 0.000001154 Vs 17:11:26 Integrated Field Strength, Average and RMS of Measurements X0 = -0.0200000 m, dx = 0.0050000 m, Nturns = 10.0 VT = 0.002946121 +- 0.000001154 Vs BL = 0.05892243 +- 0.00002308 Tm Remeasure 17:11:26 Integrated Voltage Measurement... 17:11:26 Magnet Current for MAIN, Imag = 154.9751 A Vtrans = 1.549804 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:11:34 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = 0.0050000 m VT = 0.002938314 Vs 17:11:41 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = -0.0050000 m VT = -0.002956068 Vs 17:11:42 Magnet Current for MAIN, Imag = 154.9751 A Vtrans = 1.549804 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:11:50 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = 0.0050000 m VT = 0.002936717 Vs 17:11:57 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = -0.0050000 m VT = -0.002952695 Vs 17:11:58 Magnet Current for MAIN, Imag = 154.9751 A Vtrans = 1.549804 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:12:06 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = 0.0050000 m VT = 0.002938587 Vs 17:12:13 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = -0.0050000 m VT = -0.002951382 Vs 17:12:14 Magnet Current for MAIN, Imag = 154.9752 A Vtrans = 1.549805 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:12:22 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = 0.0050000 m VT = 0.002940075 Vs 17:12:29 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = -0.0050000 m VT = -0.002952272 Vs 17:12:29 Integrated Voltage Average and RMS of Measurements X0 = -0.0200000 m, Delta X = 0.0050000 m VT = 0.002945764 +- 0.000000991 Vs 17:12:29 Integrated Field Strength, Average and RMS of Measurements X0 = -0.0200000 m, dx = 0.0050000 m, Nturns = 10.0 VT = 0.002945764 +- 0.000000991 Vs BL = 0.05891528 +- 0.00001983 Tm Remeasure 17:12:29 Integrated Voltage Measurement... 17:12:30 Magnet Current for MAIN, Imag = 154.9750 A Vtrans = 1.549803 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:12:38 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = 0.0050000 m VT = 0.002939489 Vs 17:12:45 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = -0.0050000 m VT = -0.002956008 Vs 17:12:46 Magnet Current for MAIN, Imag = 154.9750 A Vtrans = 1.549803 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:12:54 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = 0.0050000 m VT = 0.002936386 Vs 17:13:01 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = -0.0050000 m VT = -0.002956728 Vs 17:13:02 Magnet Current for MAIN, Imag = 154.9750 A Vtrans = 1.549803 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:13:10 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = 0.0050000 m VT = 0.002939579 Vs 17:13:17 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = -0.0050000 m VT = -0.002952936 Vs 17:13:18 Magnet Current for MAIN, Imag = 154.9753 A Vtrans = 1.549807 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:13:25 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = 0.0050000 m VT = 0.002938525 Vs 17:13:33 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = -0.0050000 m VT = -0.002953190 Vs 17:13:33 Integrated Voltage Average and RMS of Measurements X0 = -0.0200000 m, Delta X = 0.0050000 m VT = 0.002946605 +- 0.000000705 Vs 17:13:33 Integrated Field Strength, Average and RMS of Measurements X0 = -0.0200000 m, dx = 0.0050000 m, Nturns = 10.0 VT = 0.002946605 +- 0.000000705 Vs BL = 0.05893210 +- 0.00001410 Tm 17:13:33 Integrated Voltage Measurement... 17:13:34 Magnet Current for MAIN, Imag = 154.9750 A Vtrans = 1.549803 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:13:43 Integrated Voltage Measurement X0 = -0.0150000 m, Delta X = 0.0050000 m VT = 0.002951387 Vs 17:13:50 Integrated Voltage Measurement X0 = -0.0150000 m, Delta X = -0.0050000 m VT = -0.002971404 Vs 17:13:51 Magnet Current for MAIN, Imag = 154.9747 A Vtrans = 1.549800 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:13:58 Integrated Voltage Measurement X0 = -0.0150000 m, Delta X = 0.0050000 m VT = 0.002955213 Vs 17:14:06 Integrated Voltage Measurement X0 = -0.0150000 m, Delta X = -0.0050000 m VT = -0.002972219 Vs 17:14:07 Magnet Current for MAIN, Imag = 154.9747 A Vtrans = 1.549800 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:14:14 Integrated Voltage Measurement X0 = -0.0150000 m, Delta X = 0.0050000 m VT = 0.002955467 Vs 17:14:22 Integrated Voltage Measurement X0 = -0.0150000 m, Delta X = -0.0050000 m VT = -0.002971502 Vs 17:14:23 Magnet Current for MAIN, Imag = 154.9751 A Vtrans = 1.549804 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:14:31 Integrated Voltage Measurement X0 = -0.0150000 m, Delta X = 0.0050000 m VT = 0.002952584 Vs 17:14:38 Integrated Voltage Measurement X0 = -0.0150000 m, Delta X = -0.0050000 m VT = -0.002971112 Vs 17:14:38 Integrated Voltage Average and RMS of Measurements X0 = -0.0150000 m, Delta X = 0.0050000 m VT = 0.002962611 +- 0.000001005 Vs 17:14:38 Integrated Field Strength, Average and RMS of Measurements X0 = -0.0150000 m, dx = 0.0050000 m, Nturns = 10.0 VT = 0.002962611 +- 0.000001005 Vs BL = 0.05925222 +- 0.00002011 Tm Remeasure 17:14:38 Integrated Voltage Measurement... 17:14:39 Magnet Current for MAIN, Imag = 154.9748 A Vtrans = 1.549802 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:14:47 Integrated Voltage Measurement X0 = -0.0150000 m, Delta X = 0.0050000 m VT = 0.002946167 Vs 17:14:54 Integrated Voltage Measurement X0 = -0.0150000 m, Delta X = -0.0050000 m VT = -0.002971096 Vs 17:14:55 Magnet Current for MAIN, Imag = 154.9750 A Vtrans = 1.549803 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:15:03 Integrated Voltage Measurement X0 = -0.0150000 m, Delta X = 0.0050000 m VT = 0.002945114 Vs 17:15:10 Integrated Voltage Measurement X0 = -0.0150000 m, Delta X = -0.0050000 m VT = -0.002970864 Vs 17:15:11 Magnet Current for MAIN, Imag = 154.9751 A Vtrans = 1.549804 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:15:19 Integrated Voltage Measurement X0 = -0.0150000 m, Delta X = 0.0050000 m VT = 0.002951469 Vs 17:15:26 Integrated Voltage Measurement X0 = -0.0150000 m, Delta X = -0.0050000 m VT = -0.002967756 Vs 17:15:27 Magnet Current for MAIN, Imag = 154.9748 A Vtrans = 1.549802 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:15:34 Integrated Voltage Measurement X0 = -0.0150000 m, Delta X = 0.0050000 m VT = 0.002953389 Vs 17:15:42 Integrated Voltage Measurement X0 = -0.0150000 m, Delta X = -0.0050000 m VT = -0.002968359 Vs 17:15:42 Integrated Voltage Average and RMS of Measurements X0 = -0.0150000 m, Delta X = 0.0050000 m VT = 0.002959277 +- 0.000001088 Vs 17:15:42 Integrated Field Strength, Average and RMS of Measurements X0 = -0.0150000 m, dx = 0.0050000 m, Nturns = 10.0 VT = 0.002959277 +- 0.000001088 Vs BL = 0.05918554 +- 0.00002177 Tm 17:15:42 Integrated Voltage Measurement... 17:15:43 Magnet Current for MAIN, Imag = 154.9746 A Vtrans = 1.549799 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:15:52 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = 0.0050000 m VT = 0.002957362 Vs 17:15:59 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = -0.0050000 m VT = -0.002970680 Vs 17:16:00 Magnet Current for MAIN, Imag = 154.9748 A Vtrans = 1.549802 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:16:08 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = 0.0050000 m VT = 0.002957632 Vs 17:16:16 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = -0.0050000 m VT = -0.002970362 Vs 17:16:17 Magnet Current for MAIN, Imag = 154.9747 A Vtrans = 1.549800 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:16:25 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = 0.0050000 m VT = 0.002956359 Vs 17:16:32 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = -0.0050000 m VT = -0.002977082 Vs 17:16:33 Magnet Current for MAIN, Imag = 154.9748 A Vtrans = 1.549802 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:16:40 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = 0.0050000 m VT = 0.002957768 Vs 17:16:48 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = -0.0050000 m VT = -0.002976494 Vs 17:16:48 Integrated Voltage Average and RMS of Measurements X0 = -0.0100000 m, Delta X = 0.0050000 m VT = 0.002965467 +- 0.000001466 Vs 17:16:48 Integrated Field Strength, Average and RMS of Measurements X0 = -0.0100000 m, dx = 0.0050000 m, Nturns = 10.0 VT = 0.002965467 +- 0.000001466 Vs BL = 0.05930935 +- 0.00002931 Tm Remeasure 17:16:48 Integrated Voltage Measurement... 17:16:49 Magnet Current for MAIN, Imag = 154.9750 A Vtrans = 1.549803 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:16:56 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = 0.0050000 m VT = 0.002958692 Vs 17:17:04 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = -0.0050000 m VT = -0.002972944 Vs 17:17:05 Magnet Current for MAIN, Imag = 154.9753 A Vtrans = 1.549807 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:17:12 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = 0.0050000 m VT = 0.002956800 Vs 17:17:20 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = -0.0050000 m VT = -0.002969863 Vs 17:17:21 Magnet Current for MAIN, Imag = 154.9753 A Vtrans = 1.549807 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:17:29 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = 0.0050000 m VT = 0.002958617 Vs 17:17:36 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = -0.0050000 m VT = -0.002974220 Vs 17:17:37 Magnet Current for MAIN, Imag = 154.9750 A Vtrans = 1.549803 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:17:45 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = 0.0050000 m VT = 0.002955732 Vs 17:17:52 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = -0.0050000 m VT = -0.002978229 Vs 17:17:52 Integrated Voltage Average and RMS of Measurements X0 = -0.0100000 m, Delta X = 0.0050000 m VT = 0.002965637 +- 0.000001393 Vs 17:17:52 Integrated Field Strength, Average and RMS of Measurements X0 = -0.0100000 m, dx = 0.0050000 m, Nturns = 10.0 VT = 0.002965637 +- 0.000001393 Vs BL = 0.05931274 +- 0.00002786 Tm Remeasure 17:17:52 Integrated Voltage Measurement... 17:17:53 Magnet Current for MAIN, Imag = 154.9752 A Vtrans = 1.549805 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:18:00 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = 0.0050000 m VT = 0.002957011 Vs 17:18:08 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = -0.0050000 m VT = -0.002967192 Vs 17:18:09 Magnet Current for MAIN, Imag = 154.9748 A Vtrans = 1.549802 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:18:16 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = 0.0050000 m VT = 0.002958933 Vs 17:18:24 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = -0.0050000 m VT = -0.002972935 Vs 17:18:25 Magnet Current for MAIN, Imag = 154.9753 A Vtrans = 1.549807 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:18:32 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = 0.0050000 m VT = 0.002958310 Vs 17:18:40 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = -0.0050000 m VT = -0.002978774 Vs 17:18:41 Magnet Current for MAIN, Imag = 154.9750 A Vtrans = 1.549803 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:18:48 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = 0.0050000 m VT = 0.002958609 Vs 17:18:56 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = -0.0050000 m VT = -0.002966016 Vs 17:18:56 Integrated Voltage Average and RMS of Measurements X0 = -0.0100000 m, Delta X = 0.0050000 m VT = 0.002964722 +- 0.000002680 Vs 17:18:56 Integrated Field Strength, Average and RMS of Measurements X0 = -0.0100000 m, dx = 0.0050000 m, Nturns = 10.0 VT = 0.002964722 +- 0.000002680 Vs BL = 0.05929445 +- 0.00005360 Tm Remeasure 17:18:56 Integrated Voltage Measurement... 17:18:56 Magnet Current for MAIN, Imag = 154.9750 A Vtrans = 1.549803 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:19:04 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = 0.0050000 m VT = 0.002959375 Vs 17:19:12 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = -0.0050000 m VT = -0.002975794 Vs 17:19:12 Magnet Current for MAIN, Imag = 154.9748 A Vtrans = 1.549802 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:19:20 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = 0.0050000 m VT = 0.002961544 Vs 17:19:27 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = -0.0050000 m VT = -0.002978211 Vs 17:19:28 Magnet Current for MAIN, Imag = 154.9751 A Vtrans = 1.549804 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:19:36 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = 0.0050000 m VT = 0.002961044 Vs 17:19:44 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = -0.0050000 m VT = -0.002972659 Vs 17:19:44 Magnet Current for MAIN, Imag = 154.9747 A Vtrans = 1.549800 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:19:52 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = 0.0050000 m VT = 0.002957000 Vs 17:20:00 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = -0.0050000 m VT = -0.002975176 Vs 17:20:00 Integrated Voltage Average and RMS of Measurements X0 = -0.0100000 m, Delta X = 0.0050000 m VT = 0.002967600 +- 0.000001417 Vs 17:20:00 Integrated Field Strength, Average and RMS of Measurements X0 = -0.0100000 m, dx = 0.0050000 m, Nturns = 10.0 VT = 0.002967600 +- 0.000001417 Vs BL = 0.05935201 +- 0.00002834 Tm Remeasure 17:20:00 Integrated Voltage Measurement... 17:20:00 Magnet Current for MAIN, Imag = 154.9751 A Vtrans = 1.549804 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:20:08 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = 0.0050000 m VT = 0.002960215 Vs 17:20:15 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = -0.0050000 m VT = -0.002974533 Vs 17:20:16 Magnet Current for MAIN, Imag = 154.9752 A Vtrans = 1.549805 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:20:24 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = 0.0050000 m VT = 0.002958339 Vs 17:20:31 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = -0.0050000 m VT = -0.002973034 Vs 17:20:32 Magnet Current for MAIN, Imag = 154.9751 A Vtrans = 1.549804 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:20:40 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = 0.0050000 m VT = 0.002956540 Vs 17:20:47 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = -0.0050000 m VT = -0.002972574 Vs 17:20:48 Magnet Current for MAIN, Imag = 154.9750 A Vtrans = 1.549803 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:20:56 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = 0.0050000 m VT = 0.002955445 Vs 17:21:03 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = -0.0050000 m VT = -0.002973636 Vs 17:21:03 Integrated Voltage Average and RMS of Measurements X0 = -0.0100000 m, Delta X = 0.0050000 m VT = 0.002965540 +- 0.000001157 Vs 17:21:03 Integrated Field Strength, Average and RMS of Measurements X0 = -0.0100000 m, dx = 0.0050000 m, Nturns = 10.0 VT = 0.002965540 +- 0.000001157 Vs BL = 0.05931079 +- 0.00002313 Tm 17:21:03 Integrated Voltage Measurement... 17:21:04 Magnet Current for MAIN, Imag = 154.9750 A Vtrans = 1.549803 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:21:13 Integrated Voltage Measurement X0 = -0.0050000 m, Delta X = 0.0050000 m VT = 0.002956134 Vs 17:21:20 Integrated Voltage Measurement X0 = -0.0050000 m, Delta X = -0.0050000 m VT = -0.002974516 Vs 17:21:21 Magnet Current for MAIN, Imag = 154.9747 A Vtrans = 1.549800 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:21:29 Integrated Voltage Measurement X0 = -0.0050000 m, Delta X = 0.0050000 m VT = 0.002961540 Vs 17:21:36 Integrated Voltage Measurement X0 = -0.0050000 m, Delta X = -0.0050000 m VT = -0.002973974 Vs 17:21:37 Magnet Current for MAIN, Imag = 154.9748 A Vtrans = 1.549802 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:21:45 Integrated Voltage Measurement X0 = -0.0050000 m, Delta X = 0.0050000 m VT = 0.002961297 Vs 17:21:52 Integrated Voltage Measurement X0 = -0.0050000 m, Delta X = -0.0050000 m VT = -0.002973224 Vs 17:21:53 Magnet Current for MAIN, Imag = 154.9752 A Vtrans = 1.549805 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:22:01 Integrated Voltage Measurement X0 = -0.0050000 m, Delta X = 0.0050000 m VT = 0.002958399 Vs 17:22:08 Integrated Voltage Measurement X0 = -0.0050000 m, Delta X = -0.0050000 m VT = -0.002973739 Vs 17:22:08 Integrated Voltage Average and RMS of Measurements X0 = -0.0050000 m, Delta X = 0.0050000 m VT = 0.002966603 +- 0.000000959 Vs 17:22:08 Integrated Field Strength, Average and RMS of Measurements X0 = -0.0050000 m, dx = 0.0050000 m, Nturns = 10.0 VT = 0.002966603 +- 0.000000959 Vs BL = 0.05933206 +- 0.00001918 Tm Remeasure 17:22:08 Integrated Voltage Measurement... 17:22:09 Magnet Current for MAIN, Imag = 154.9752 A Vtrans = 1.549805 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:22:17 Integrated Voltage Measurement X0 = -0.0050000 m, Delta X = 0.0050000 m VT = 0.002962873 Vs 17:22:24 Integrated Voltage Measurement X0 = -0.0050000 m, Delta X = -0.0050000 m VT = -0.002973207 Vs 17:22:25 Magnet Current for MAIN, Imag = 154.9750 A Vtrans = 1.549803 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:22:32 Integrated Voltage Measurement X0 = -0.0050000 m, Delta X = 0.0050000 m VT = 0.002959090 Vs 17:22:40 Integrated Voltage Measurement X0 = -0.0050000 m, Delta X = -0.0050000 m VT = -0.002972138 Vs 17:22:40 Magnet Current for MAIN, Imag = 154.9750 A Vtrans = 1.549803 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:22:48 Integrated Voltage Measurement X0 = -0.0050000 m, Delta X = 0.0050000 m VT = 0.002960833 Vs 17:22:56 Integrated Voltage Measurement X0 = -0.0050000 m, Delta X = -0.0050000 m VT = -0.002975522 Vs 17:22:57 Magnet Current for MAIN, Imag = 154.9751 A Vtrans = 1.549804 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:23:04 Integrated Voltage Measurement X0 = -0.0050000 m, Delta X = 0.0050000 m VT = 0.002957483 Vs 17:23:12 Integrated Voltage Measurement X0 = -0.0050000 m, Delta X = -0.0050000 m VT = -0.002974851 Vs 17:23:12 Integrated Voltage Average and RMS of Measurements X0 = -0.0050000 m, Delta X = 0.0050000 m VT = 0.002967000 +- 0.000001127 Vs 17:23:12 Integrated Field Strength, Average and RMS of Measurements X0 = -0.0050000 m, dx = 0.0050000 m, Nturns = 10.0 VT = 0.002967000 +- 0.000001127 Vs BL = 0.05933999 +- 0.00002255 Tm 17:23:12 Integrated Voltage Measurement... 17:23:12 Magnet Current for MAIN, Imag = 154.9748 A Vtrans = 1.549802 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:23:21 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0050000 m VT = 0.002960417 Vs 17:23:29 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0050000 m VT = -0.002974119 Vs 17:23:29 Magnet Current for MAIN, Imag = 154.9750 A Vtrans = 1.549803 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:23:37 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0050000 m VT = 0.002959471 Vs 17:23:45 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0050000 m VT = -0.002973959 Vs 17:23:45 Magnet Current for MAIN, Imag = 154.9750 A Vtrans = 1.549803 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:23:53 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0050000 m VT = 0.002959763 Vs 17:24:00 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0050000 m VT = -0.002975038 Vs 17:24:01 Magnet Current for MAIN, Imag = 154.9750 A Vtrans = 1.549803 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:24:09 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0050000 m VT = 0.002959584 Vs 17:24:16 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0050000 m VT = -0.002976180 Vs 17:24:16 Integrated Voltage Average and RMS of Measurements X0 = 0.0000000 m, Delta X = 0.0050000 m VT = 0.002967316 +- 0.000000416 Vs 17:24:16 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0000000 m, dx = 0.0050000 m, Nturns = 10.0 VT = 0.002967316 +- 0.000000416 Vs BL = 0.05934633 +- 0.00000831 Tm 17:24:16 Integrated Voltage Measurement... 17:24:17 Magnet Current for MAIN, Imag = 154.9751 A Vtrans = 1.549804 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:24:26 Integrated Voltage Measurement X0 = 0.0050000 m, Delta X = 0.0050000 m VT = 0.002954348 Vs 17:24:33 Integrated Voltage Measurement X0 = 0.0050000 m, Delta X = -0.0050000 m VT = -0.002977166 Vs 17:24:34 Magnet Current for MAIN, Imag = 154.9753 A Vtrans = 1.549807 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:24:41 Integrated Voltage Measurement X0 = 0.0050000 m, Delta X = 0.0050000 m VT = 0.002953293 Vs 17:24:49 Integrated Voltage Measurement X0 = 0.0050000 m, Delta X = -0.0050000 m VT = -0.002976073 Vs 17:24:50 Magnet Current for MAIN, Imag = 154.9747 A Vtrans = 1.549800 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:24:57 Integrated Voltage Measurement X0 = 0.0050000 m, Delta X = 0.0050000 m VT = 0.002957725 Vs 17:25:05 Integrated Voltage Measurement X0 = 0.0050000 m, Delta X = -0.0050000 m VT = -0.002973585 Vs 17:25:06 Magnet Current for MAIN, Imag = 154.9748 A Vtrans = 1.549802 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:25:13 Integrated Voltage Measurement X0 = 0.0050000 m, Delta X = 0.0050000 m VT = 0.002960701 Vs 17:25:21 Integrated Voltage Measurement X0 = 0.0050000 m, Delta X = -0.0050000 m VT = -0.002973308 Vs 17:25:21 Integrated Voltage Average and RMS of Measurements X0 = 0.0050000 m, Delta X = 0.0050000 m VT = 0.002965775 +- 0.000000824 Vs 17:25:21 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0050000 m, dx = 0.0050000 m, Nturns = 10.0 VT = 0.002965775 +- 0.000000824 Vs BL = 0.05931549 +- 0.00001649 Tm Remeasure 17:25:21 Integrated Voltage Measurement... 17:25:22 Magnet Current for MAIN, Imag = 154.9747 A Vtrans = 1.549800 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:25:29 Integrated Voltage Measurement X0 = 0.0050000 m, Delta X = 0.0050000 m VT = 0.002958438 Vs 17:25:37 Integrated Voltage Measurement X0 = 0.0050000 m, Delta X = -0.0050000 m VT = -0.002976368 Vs 17:25:38 Magnet Current for MAIN, Imag = 154.9752 A Vtrans = 1.549805 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:25:45 Integrated Voltage Measurement X0 = 0.0050000 m, Delta X = 0.0050000 m VT = 0.002960164 Vs 17:25:53 Integrated Voltage Measurement X0 = 0.0050000 m, Delta X = -0.0050000 m VT = -0.002974376 Vs 17:25:54 Magnet Current for MAIN, Imag = 154.9750 A Vtrans = 1.549803 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:26:01 Integrated Voltage Measurement X0 = 0.0050000 m, Delta X = 0.0050000 m VT = 0.002959693 Vs 17:26:09 Integrated Voltage Measurement X0 = 0.0050000 m, Delta X = -0.0050000 m VT = -0.002973357 Vs 17:26:09 Magnet Current for MAIN, Imag = 154.9751 A Vtrans = 1.549804 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:26:17 Integrated Voltage Measurement X0 = 0.0050000 m, Delta X = 0.0050000 m VT = 0.002960953 Vs 17:26:25 Integrated Voltage Measurement X0 = 0.0050000 m, Delta X = -0.0050000 m VT = -0.002976958 Vs 17:26:25 Integrated Voltage Average and RMS of Measurements X0 = 0.0050000 m, Delta X = 0.0050000 m VT = 0.002967538 +- 0.000000884 Vs 17:26:25 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0050000 m, dx = 0.0050000 m, Nturns = 10.0 VT = 0.002967538 +- 0.000000884 Vs BL = 0.05935077 +- 0.00001768 Tm 17:26:25 Integrated Voltage Measurement... 17:26:25 Magnet Current for MAIN, Imag = 154.9751 A Vtrans = 1.549804 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:26:34 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = 0.0050000 m VT = 0.002948316 Vs 17:26:42 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = -0.0050000 m VT = -0.002978084 Vs 17:26:42 Magnet Current for MAIN, Imag = 154.9750 A Vtrans = 1.549803 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:26:50 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = 0.0050000 m VT = 0.002957071 Vs 17:26:58 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = -0.0050000 m VT = -0.002969281 Vs 17:26:58 Magnet Current for MAIN, Imag = 154.9750 A Vtrans = 1.549803 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:27:06 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = 0.0050000 m VT = 0.002971253 Vs 17:27:14 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = -0.0050000 m VT = -0.002973627 Vs 17:27:14 Magnet Current for MAIN, Imag = 154.9747 A Vtrans = 1.549800 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:27:22 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = 0.0050000 m VT = 0.002956608 Vs 17:27:30 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = -0.0050000 m VT = -0.002976567 Vs 17:27:30 Integrated Voltage Average and RMS of Measurements X0 = 0.0100000 m, Delta X = 0.0050000 m VT = 0.002966351 +- 0.000003780 Vs 17:27:30 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0100000 m, dx = 0.0050000 m, Nturns = 10.0 VT = 0.002966351 +- 0.000003780 Vs BL = 0.05932702 +- 0.00007559 Tm Remeasure 17:27:30 Integrated Voltage Measurement... 17:27:30 Magnet Current for MAIN, Imag = 154.9752 A Vtrans = 1.549805 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:27:38 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = 0.0050000 m VT = 0.002953126 Vs 17:27:46 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = -0.0050000 m VT = -0.002970253 Vs 17:27:46 Magnet Current for MAIN, Imag = 154.9751 A Vtrans = 1.549804 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:27:54 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = 0.0050000 m VT = 0.002944465 Vs 17:28:02 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = -0.0050000 m VT = -0.002973387 Vs 17:28:02 Magnet Current for MAIN, Imag = 154.9746 A Vtrans = 1.549799 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:28:10 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = 0.0050000 m VT = 0.002961747 Vs 17:28:18 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = -0.0050000 m VT = -0.002969106 Vs 17:28:18 Magnet Current for MAIN, Imag = 154.9750 A Vtrans = 1.549803 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:28:26 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = 0.0050000 m VT = 0.002932304 Vs 17:28:34 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = -0.0050000 m VT = -0.002975127 Vs 17:28:34 Integrated Voltage Average and RMS of Measurements X0 = 0.0100000 m, Delta X = 0.0050000 m VT = 0.002959939 +- 0.000004270 Vs 17:28:34 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0100000 m, dx = 0.0050000 m, Nturns = 10.0 VT = 0.002959939 +- 0.000004270 Vs BL = 0.05919879 +- 0.00008540 Tm Remeasure 17:28:34 Integrated Voltage Measurement... 17:28:34 Magnet Current for MAIN, Imag = 154.9753 A Vtrans = 1.549807 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:28:42 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = 0.0050000 m VT = 0.002961067 Vs 17:28:50 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = -0.0050000 m VT = -0.002973191 Vs 17:28:50 Magnet Current for MAIN, Imag = 154.9751 A Vtrans = 1.549804 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:28:58 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = 0.0050000 m VT = 0.002965174 Vs 17:29:06 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = -0.0050000 m VT = -0.002975084 Vs 17:29:06 Magnet Current for MAIN, Imag = 154.9748 A Vtrans = 1.549802 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:29:14 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = 0.0050000 m VT = 0.002950050 Vs 17:29:21 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = -0.0050000 m VT = -0.002978063 Vs 17:29:22 Magnet Current for MAIN, Imag = 154.9750 A Vtrans = 1.549803 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:29:30 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = 0.0050000 m VT = 0.002943289 Vs 17:29:37 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = -0.0050000 m VT = -0.002973400 Vs 17:29:37 Integrated Voltage Average and RMS of Measurements X0 = 0.0100000 m, Delta X = 0.0050000 m VT = 0.002964915 +- 0.000004359 Vs 17:29:37 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0100000 m, dx = 0.0050000 m, Nturns = 10.0 VT = 0.002964915 +- 0.000004359 Vs BL = 0.05929830 +- 0.00008717 Tm Remeasure 17:29:37 Integrated Voltage Measurement... 17:29:38 Magnet Current for MAIN, Imag = 154.9751 A Vtrans = 1.549804 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:29:46 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = 0.0050000 m VT = 0.002948051 Vs 17:29:53 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = -0.0050000 m VT = -0.002968527 Vs 17:29:54 Magnet Current for MAIN, Imag = 154.9753 A Vtrans = 1.549807 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:30:02 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = 0.0050000 m VT = 0.002968488 Vs 17:30:09 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = -0.0050000 m VT = -0.002973933 Vs 17:30:10 Magnet Current for MAIN, Imag = 154.9750 A Vtrans = 1.549803 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:30:17 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = 0.0050000 m VT = 0.002976182 Vs 17:30:25 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = -0.0050000 m VT = -0.002975637 Vs 17:30:26 Magnet Current for MAIN, Imag = 154.9747 A Vtrans = 1.549800 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:30:33 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = 0.0050000 m VT = 0.002952159 Vs 17:30:41 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = -0.0050000 m VT = -0.002977711 Vs 17:30:41 Integrated Voltage Average and RMS of Measurements X0 = 0.0100000 m, Delta X = 0.0050000 m VT = 0.002967586 +- 0.000006631 Vs 17:30:41 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0100000 m, dx = 0.0050000 m, Nturns = 10.0 VT = 0.002967586 +- 0.000006631 Vs BL = 0.05935172 +- 0.00013262 Tm Remeasure 17:30:41 Integrated Voltage Measurement... 17:30:41 Magnet Current for MAIN, Imag = 154.9752 A Vtrans = 1.549805 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:30:49 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = 0.0050000 m VT = 0.002965133 Vs 17:30:57 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = -0.0050000 m VT = -0.002976522 Vs 17:30:57 Magnet Current for MAIN, Imag = 154.9752 A Vtrans = 1.549805 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:31:05 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = 0.0050000 m VT = 0.002943042 Vs 17:31:13 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = -0.0050000 m VT = -0.002973426 Vs 17:31:13 Magnet Current for MAIN, Imag = 154.9747 A Vtrans = 1.549800 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:31:21 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = 0.0050000 m VT = 0.002939512 Vs 17:31:29 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = -0.0050000 m VT = -0.002973839 Vs 17:31:29 Magnet Current for MAIN, Imag = 154.9751 A Vtrans = 1.549804 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:31:37 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = 0.0050000 m VT = 0.002946236 Vs 17:31:45 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = -0.0050000 m VT = -0.002976724 Vs 17:31:45 Integrated Voltage Average and RMS of Measurements X0 = 0.0100000 m, Delta X = 0.0050000 m VT = 0.002961804 +- 0.000005490 Vs 17:31:45 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0100000 m, dx = 0.0050000 m, Nturns = 10.0 VT = 0.002961804 +- 0.000005490 Vs BL = 0.05923608 +- 0.00010981 Tm 17:31:45 Integrated Voltage Measurement... 17:31:45 Magnet Current for MAIN, Imag = 154.9753 A Vtrans = 1.549807 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:31:54 Integrated Voltage Measurement X0 = 0.0150000 m, Delta X = 0.0050000 m VT = 0.002952950 Vs 17:32:02 Integrated Voltage Measurement X0 = 0.0150000 m, Delta X = -0.0050000 m VT = -0.002976469 Vs 17:32:02 Magnet Current for MAIN, Imag = 154.9752 A Vtrans = 1.549805 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:32:10 Integrated Voltage Measurement X0 = 0.0150000 m, Delta X = 0.0050000 m VT = 0.002955178 Vs 17:32:18 Integrated Voltage Measurement X0 = 0.0150000 m, Delta X = -0.0050000 m VT = -0.002981508 Vs 17:32:18 Magnet Current for MAIN, Imag = 154.9750 A Vtrans = 1.549803 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:32:26 Integrated Voltage Measurement X0 = 0.0150000 m, Delta X = 0.0050000 m VT = 0.002954063 Vs 17:32:33 Integrated Voltage Measurement X0 = 0.0150000 m, Delta X = -0.0050000 m VT = -0.002950528 Vs 17:32:34 Magnet Current for MAIN, Imag = 154.9747 A Vtrans = 1.549800 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:32:42 Integrated Voltage Measurement X0 = 0.0150000 m, Delta X = 0.0050000 m VT = 0.002947750 Vs 17:32:49 Integrated Voltage Measurement X0 = 0.0150000 m, Delta X = -0.0050000 m VT = -0.002975525 Vs 17:32:49 Integrated Voltage Average and RMS of Measurements X0 = 0.0150000 m, Delta X = 0.0050000 m VT = 0.002961746 +- 0.000005950 Vs 17:32:49 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0150000 m, dx = 0.0050000 m, Nturns = 10.0 VT = 0.002961746 +- 0.000005950 Vs BL = 0.05923493 +- 0.00011900 Tm Remeasure 17:32:49 Integrated Voltage Measurement... 17:32:50 Magnet Current for MAIN, Imag = 154.9747 A Vtrans = 1.549800 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:32:58 Integrated Voltage Measurement X0 = 0.0150000 m, Delta X = 0.0050000 m VT = 0.002952540 Vs 17:33:05 Integrated Voltage Measurement X0 = 0.0150000 m, Delta X = -0.0050000 m VT = -0.002972699 Vs 17:33:06 Magnet Current for MAIN, Imag = 154.9752 A Vtrans = 1.549805 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:33:14 Integrated Voltage Measurement X0 = 0.0150000 m, Delta X = 0.0050000 m VT = 0.002954786 Vs 17:33:21 Integrated Voltage Measurement X0 = 0.0150000 m, Delta X = -0.0050000 m VT = -0.002995301 Vs 17:33:22 Magnet Current for MAIN, Imag = 154.9755 A Vtrans = 1.549808 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:33:30 Integrated Voltage Measurement X0 = 0.0150000 m, Delta X = 0.0050000 m VT = 0.002954339 Vs 17:33:37 Integrated Voltage Measurement X0 = 0.0150000 m, Delta X = -0.0050000 m VT = -0.002981419 Vs 17:33:38 Magnet Current for MAIN, Imag = 154.9750 A Vtrans = 1.549803 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:33:46 Integrated Voltage Measurement X0 = 0.0150000 m, Delta X = 0.0050000 m VT = 0.002955425 Vs 17:33:53 Integrated Voltage Measurement X0 = 0.0150000 m, Delta X = -0.0050000 m VT = -0.002984546 Vs 17:33:53 Integrated Voltage Average and RMS of Measurements X0 = 0.0150000 m, Delta X = 0.0050000 m VT = 0.002968882 +- 0.000004455 Vs 17:33:53 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0150000 m, dx = 0.0050000 m, Nturns = 10.0 VT = 0.002968882 +- 0.000004455 Vs BL = 0.05937764 +- 0.00008911 Tm Remeasure 17:33:53 Integrated Voltage Measurement... 17:33:54 Magnet Current for MAIN, Imag = 154.9752 A Vtrans = 1.549805 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:34:02 Integrated Voltage Measurement X0 = 0.0150000 m, Delta X = 0.0050000 m VT = 0.002946003 Vs 17:34:09 Integrated Voltage Measurement X0 = 0.0150000 m, Delta X = -0.0050000 m VT = -0.002964992 Vs 17:34:10 Magnet Current for MAIN, Imag = 154.9750 A Vtrans = 1.549803 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:34:18 Integrated Voltage Measurement X0 = 0.0150000 m, Delta X = 0.0050000 m VT = 0.002948871 Vs 17:34:25 Integrated Voltage Measurement X0 = 0.0150000 m, Delta X = -0.0050000 m VT = -0.002963898 Vs 17:34:26 Magnet Current for MAIN, Imag = 154.9751 A Vtrans = 1.549804 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:34:33 Integrated Voltage Measurement X0 = 0.0150000 m, Delta X = 0.0050000 m VT = 0.002953206 Vs 17:34:41 Integrated Voltage Measurement X0 = 0.0150000 m, Delta X = -0.0050000 m VT = -0.002978938 Vs 17:34:42 Magnet Current for MAIN, Imag = 154.9747 A Vtrans = 1.549800 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:34:49 Integrated Voltage Measurement X0 = 0.0150000 m, Delta X = 0.0050000 m VT = 0.002958284 Vs 17:34:57 Integrated Voltage Measurement X0 = 0.0150000 m, Delta X = -0.0050000 m VT = -0.002958934 Vs 17:34:57 Integrated Voltage Average and RMS of Measurements X0 = 0.0150000 m, Delta X = 0.0050000 m VT = 0.002959141 +- 0.000004159 Vs 17:34:57 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0150000 m, dx = 0.0050000 m, Nturns = 10.0 VT = 0.002959141 +- 0.000004159 Vs BL = 0.05918282 +- 0.00008318 Tm Remeasure 17:34:57 Integrated Voltage Measurement... 17:34:58 Magnet Current for MAIN, Imag = 154.9750 A Vtrans = 1.549803 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:35:05 Integrated Voltage Measurement X0 = 0.0150000 m, Delta X = 0.0050000 m VT = 0.002948855 Vs 17:35:13 Integrated Voltage Measurement X0 = 0.0150000 m, Delta X = -0.0050000 m VT = -0.002985481 Vs 17:35:13 Magnet Current for MAIN, Imag = 154.9750 A Vtrans = 1.549803 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:35:21 Integrated Voltage Measurement X0 = 0.0150000 m, Delta X = 0.0050000 m VT = 0.002954522 Vs 17:35:29 Integrated Voltage Measurement X0 = 0.0150000 m, Delta X = -0.0050000 m VT = -0.002969513 Vs 17:35:30 Magnet Current for MAIN, Imag = 154.9750 A Vtrans = 1.549803 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:35:37 Integrated Voltage Measurement X0 = 0.0150000 m, Delta X = 0.0050000 m VT = 0.002950569 Vs 17:35:45 Integrated Voltage Measurement X0 = 0.0150000 m, Delta X = -0.0050000 m VT = -0.002960149 Vs 17:35:45 Magnet Current for MAIN, Imag = 154.9746 A Vtrans = 1.549799 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:35:53 Integrated Voltage Measurement X0 = 0.0150000 m, Delta X = 0.0050000 m VT = 0.002946456 Vs 17:36:01 Integrated Voltage Measurement X0 = 0.0150000 m, Delta X = -0.0050000 m VT = -0.002982681 Vs 17:36:01 Integrated Voltage Average and RMS of Measurements X0 = 0.0150000 m, Delta X = 0.0050000 m VT = 0.002962278 +- 0.000004390 Vs 17:36:01 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0150000 m, dx = 0.0050000 m, Nturns = 10.0 VT = 0.002962278 +- 0.000004390 Vs BL = 0.05924556 +- 0.00008780 Tm Remeasure 17:36:01 Integrated Voltage Measurement... 17:36:01 Magnet Current for MAIN, Imag = 154.9753 A Vtrans = 1.549807 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:36:09 Integrated Voltage Measurement X0 = 0.0150000 m, Delta X = 0.0050000 m VT = 0.002952226 Vs 17:36:17 Integrated Voltage Measurement X0 = 0.0150000 m, Delta X = -0.0050000 m VT = -0.002976644 Vs 17:36:17 Magnet Current for MAIN, Imag = 154.9750 A Vtrans = 1.549803 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:36:25 Integrated Voltage Measurement X0 = 0.0150000 m, Delta X = 0.0050000 m VT = 0.002950582 Vs 17:36:32 Integrated Voltage Measurement X0 = 0.0150000 m, Delta X = -0.0050000 m VT = -0.002957247 Vs 17:36:33 Magnet Current for MAIN, Imag = 154.9753 A Vtrans = 1.549807 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:36:41 Integrated Voltage Measurement X0 = 0.0150000 m, Delta X = 0.0050000 m VT = 0.002954074 Vs 17:36:48 Integrated Voltage Measurement X0 = 0.0150000 m, Delta X = -0.0050000 m VT = -0.002963591 Vs 17:36:49 Magnet Current for MAIN, Imag = 154.9750 A Vtrans = 1.549803 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:36:57 Integrated Voltage Measurement X0 = 0.0150000 m, Delta X = 0.0050000 m VT = 0.002947879 Vs 17:37:04 Integrated Voltage Measurement X0 = 0.0150000 m, Delta X = -0.0050000 m VT = -0.002977646 Vs 17:37:04 Integrated Voltage Average and RMS of Measurements X0 = 0.0150000 m, Delta X = 0.0050000 m VT = 0.002959986 +- 0.000004053 Vs 17:37:04 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0150000 m, dx = 0.0050000 m, Nturns = 10.0 VT = 0.002959986 +- 0.000004053 Vs BL = 0.05919972 +- 0.00008106 Tm 17:37:04 Integrated Voltage Measurement... 17:37:05 Magnet Current for MAIN, Imag = 154.9752 A Vtrans = 1.549805 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:37:14 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = 0.0050000 m VT = 0.002933900 Vs 17:37:21 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = -0.0050000 m VT = -0.002954336 Vs 17:37:22 Magnet Current for MAIN, Imag = 154.9750 A Vtrans = 1.549803 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:37:30 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = 0.0050000 m VT = 0.002940251 Vs 17:37:37 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = -0.0050000 m VT = -0.002954563 Vs 17:37:38 Magnet Current for MAIN, Imag = 154.9748 A Vtrans = 1.549802 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:37:46 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = 0.0050000 m VT = 0.002941689 Vs 17:37:53 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = -0.0050000 m VT = -0.002955000 Vs 17:37:54 Magnet Current for MAIN, Imag = 154.9750 A Vtrans = 1.549803 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:38:02 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = 0.0050000 m VT = 0.002933252 Vs 17:38:09 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = -0.0050000 m VT = -0.002955624 Vs 17:38:09 Integrated Voltage Average and RMS of Measurements X0 = 0.0200000 m, Delta X = 0.0050000 m VT = 0.002946077 +- 0.000001833 Vs 17:38:09 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0200000 m, dx = 0.0050000 m, Nturns = 10.0 VT = 0.002946077 +- 0.000001833 Vs BL = 0.05892154 +- 0.00003665 Tm Remeasure 17:38:09 Integrated Voltage Measurement... 17:38:10 Magnet Current for MAIN, Imag = 154.9750 A Vtrans = 1.549803 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:38:18 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = 0.0050000 m VT = 0.002938074 Vs 17:38:25 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = -0.0050000 m VT = -0.002949407 Vs 17:38:26 Magnet Current for MAIN, Imag = 154.9750 A Vtrans = 1.549803 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:38:34 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = 0.0050000 m VT = 0.002941017 Vs 17:38:41 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = -0.0050000 m VT = -0.002948502 Vs 17:38:42 Magnet Current for MAIN, Imag = 154.9752 A Vtrans = 1.549805 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:38:50 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = 0.0050000 m VT = 0.002938109 Vs 17:38:57 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = -0.0050000 m VT = -0.002948976 Vs 17:38:58 Magnet Current for MAIN, Imag = 154.9753 A Vtrans = 1.549807 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:39:05 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = 0.0050000 m VT = 0.002939499 Vs 17:39:13 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = -0.0050000 m VT = -0.002951368 Vs 17:39:13 Integrated Voltage Average and RMS of Measurements X0 = 0.0200000 m, Delta X = 0.0050000 m VT = 0.002944369 +- 0.000000769 Vs 17:39:13 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0200000 m, dx = 0.0050000 m, Nturns = 10.0 VT = 0.002944369 +- 0.000000769 Vs BL = 0.05888738 +- 0.00001537 Tm 17:39:13 Integrated Voltage Measurement... 17:39:14 Magnet Current for MAIN, Imag = 154.9751 A Vtrans = 1.549804 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:39:22 Integrated Voltage Measurement X0 = 0.0250000 m, Delta X = 0.0050000 m VT = 0.002902117 Vs 17:39:30 Integrated Voltage Measurement X0 = 0.0250000 m, Delta X = -0.0050000 m VT = -0.002911953 Vs 17:39:31 Magnet Current for MAIN, Imag = 154.9751 A Vtrans = 1.549804 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:39:38 Integrated Voltage Measurement X0 = 0.0250000 m, Delta X = 0.0050000 m VT = 0.002904141 Vs 17:39:46 Integrated Voltage Measurement X0 = 0.0250000 m, Delta X = -0.0050000 m VT = -0.002909524 Vs 17:39:47 Magnet Current for MAIN, Imag = 154.9753 A Vtrans = 1.549807 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:39:54 Integrated Voltage Measurement X0 = 0.0250000 m, Delta X = 0.0050000 m VT = 0.002900381 Vs 17:40:02 Integrated Voltage Measurement X0 = 0.0250000 m, Delta X = -0.0050000 m VT = -0.002914959 Vs 17:40:03 Magnet Current for MAIN, Imag = 154.9747 A Vtrans = 1.549800 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:40:10 Integrated Voltage Measurement X0 = 0.0250000 m, Delta X = 0.0050000 m VT = 0.002902686 Vs 17:40:18 Integrated Voltage Measurement X0 = 0.0250000 m, Delta X = -0.0050000 m VT = -0.002914433 Vs 17:40:18 Integrated Voltage Average and RMS of Measurements X0 = 0.0250000 m, Delta X = 0.0050000 m VT = 0.002907524 +- 0.000000673 Vs 17:40:18 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0250000 m, dx = 0.0050000 m, Nturns = 10.0 VT = 0.002907524 +- 0.000000673 Vs BL = 0.05815049 +- 0.00001345 Tm 17:40:18 Integrated Voltage Measurement... 17:40:19 Magnet Current for MAIN, Imag = 154.9750 A Vtrans = 1.549803 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:40:27 Integrated Voltage Measurement X0 = 0.0300000 m, Delta X = 0.0050000 m VT = 0.002821471 Vs 17:40:35 Integrated Voltage Measurement X0 = 0.0300000 m, Delta X = -0.0050000 m VT = -0.002833079 Vs 17:40:36 Magnet Current for MAIN, Imag = 154.9750 A Vtrans = 1.549803 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:40:43 Integrated Voltage Measurement X0 = 0.0300000 m, Delta X = 0.0050000 m VT = 0.002819469 Vs 17:40:51 Integrated Voltage Measurement X0 = 0.0300000 m, Delta X = -0.0050000 m VT = -0.002835963 Vs 17:40:52 Magnet Current for MAIN, Imag = 154.9755 A Vtrans = 1.549808 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:40:59 Integrated Voltage Measurement X0 = 0.0300000 m, Delta X = 0.0050000 m VT = 0.002821016 Vs 17:41:07 Integrated Voltage Measurement X0 = 0.0300000 m, Delta X = -0.0050000 m VT = -0.002838561 Vs 17:41:07 Magnet Current for MAIN, Imag = 154.9750 A Vtrans = 1.549803 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:41:15 Integrated Voltage Measurement X0 = 0.0300000 m, Delta X = 0.0050000 m VT = 0.002818662 Vs 17:41:23 Integrated Voltage Measurement X0 = 0.0300000 m, Delta X = -0.0050000 m VT = -0.002831400 Vs 17:41:23 Integrated Voltage Average and RMS of Measurements X0 = 0.0300000 m, Delta X = 0.0050000 m VT = 0.002827453 +- 0.000001690 Vs 17:41:23 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0300000 m, dx = 0.0050000 m, Nturns = 10.0 VT = 0.002827453 +- 0.000001690 Vs BL = 0.05654905 +- 0.00003380 Tm Remeasure 17:41:23 Integrated Voltage Measurement... 17:41:23 Magnet Current for MAIN, Imag = 154.9747 A Vtrans = 1.549800 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:41:31 Integrated Voltage Measurement X0 = 0.0300000 m, Delta X = 0.0050000 m VT = 0.002822254 Vs 17:41:39 Integrated Voltage Measurement X0 = 0.0300000 m, Delta X = -0.0050000 m VT = -0.002836105 Vs 17:41:39 Magnet Current for MAIN, Imag = 154.9748 A Vtrans = 1.549802 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:41:47 Integrated Voltage Measurement X0 = 0.0300000 m, Delta X = 0.0050000 m VT = 0.002816601 Vs 17:41:54 Integrated Voltage Measurement X0 = 0.0300000 m, Delta X = -0.0050000 m VT = -0.002835577 Vs 17:41:55 Magnet Current for MAIN, Imag = 154.9753 A Vtrans = 1.549807 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:42:03 Integrated Voltage Measurement X0 = 0.0300000 m, Delta X = 0.0050000 m VT = 0.002824764 Vs 17:42:10 Integrated Voltage Measurement X0 = 0.0300000 m, Delta X = -0.0050000 m VT = -0.002838837 Vs 17:42:11 Magnet Current for MAIN, Imag = 154.9753 A Vtrans = 1.549807 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:42:18 Integrated Voltage Measurement X0 = 0.0300000 m, Delta X = 0.0050000 m VT = 0.002821390 Vs 17:42:26 Integrated Voltage Measurement X0 = 0.0300000 m, Delta X = -0.0050000 m VT = -0.002839392 Vs 17:42:26 Integrated Voltage Average and RMS of Measurements X0 = 0.0300000 m, Delta X = 0.0050000 m VT = 0.002829365 +- 0.000002107 Vs 17:42:26 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0300000 m, dx = 0.0050000 m, Nturns = 10.0 VT = 0.002829365 +- 0.000002107 Vs BL = 0.05658730 +- 0.00004213 Tm Remeasure 17:42:26 Integrated Voltage Measurement... 17:42:27 Magnet Current for MAIN, Imag = 154.9751 A Vtrans = 1.549804 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:42:34 Integrated Voltage Measurement X0 = 0.0300000 m, Delta X = 0.0050000 m VT = 0.002816165 Vs 17:42:42 Integrated Voltage Measurement X0 = 0.0300000 m, Delta X = -0.0050000 m VT = -0.002839888 Vs 17:42:43 Magnet Current for MAIN, Imag = 154.9747 A Vtrans = 1.549800 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:42:50 Integrated Voltage Measurement X0 = 0.0300000 m, Delta X = 0.0050000 m VT = 0.002815545 Vs 17:42:58 Integrated Voltage Measurement X0 = 0.0300000 m, Delta X = -0.0050000 m VT = -0.002832122 Vs 17:42:59 Magnet Current for MAIN, Imag = 154.9752 A Vtrans = 1.549805 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:43:06 Integrated Voltage Measurement X0 = 0.0300000 m, Delta X = 0.0050000 m VT = 0.002818613 Vs 17:43:14 Integrated Voltage Measurement X0 = 0.0300000 m, Delta X = -0.0050000 m VT = -0.002834888 Vs 17:43:14 Magnet Current for MAIN, Imag = 154.9748 A Vtrans = 1.549802 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:43:22 Integrated Voltage Measurement X0 = 0.0300000 m, Delta X = 0.0050000 m VT = 0.002818515 Vs 17:43:29 Integrated Voltage Measurement X0 = 0.0300000 m, Delta X = -0.0050000 m VT = -0.002833772 Vs 17:43:29 Integrated Voltage Average and RMS of Measurements X0 = 0.0300000 m, Delta X = 0.0050000 m VT = 0.002826189 +- 0.000001520 Vs 17:43:29 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0300000 m, dx = 0.0050000 m, Nturns = 10.0 VT = 0.002826189 +- 0.000001520 Vs BL = 0.05652377 +- 0.00003040 Tm Remeasure 17:43:29 Integrated Voltage Measurement... 17:43:30 Magnet Current for MAIN, Imag = 154.9751 A Vtrans = 1.549804 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:43:38 Integrated Voltage Measurement X0 = 0.0300000 m, Delta X = 0.0050000 m VT = 0.002822062 Vs 17:43:46 Integrated Voltage Measurement X0 = 0.0300000 m, Delta X = -0.0050000 m VT = -0.002833086 Vs 17:43:46 Magnet Current for MAIN, Imag = 154.9750 A Vtrans = 1.549803 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:43:54 Integrated Voltage Measurement X0 = 0.0300000 m, Delta X = 0.0050000 m VT = 0.002822704 Vs 17:44:02 Integrated Voltage Measurement X0 = 0.0300000 m, Delta X = -0.0050000 m VT = -0.002835750 Vs 17:44:02 Magnet Current for MAIN, Imag = 154.9748 A Vtrans = 1.549802 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:44:10 Integrated Voltage Measurement X0 = 0.0300000 m, Delta X = 0.0050000 m VT = 0.002818283 Vs 17:44:18 Integrated Voltage Measurement X0 = 0.0300000 m, Delta X = -0.0050000 m VT = -0.002834073 Vs 17:44:18 Magnet Current for MAIN, Imag = 154.9752 A Vtrans = 1.549805 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:44:26 Integrated Voltage Measurement X0 = 0.0300000 m, Delta X = 0.0050000 m VT = 0.002821374 Vs 17:44:33 Integrated Voltage Measurement X0 = 0.0300000 m, Delta X = -0.0050000 m VT = -0.002836057 Vs 17:44:33 Integrated Voltage Average and RMS of Measurements X0 = 0.0300000 m, Delta X = 0.0050000 m VT = 0.002827924 +- 0.000001172 Vs 17:44:34 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0300000 m, dx = 0.0050000 m, Nturns = 10.0 VT = 0.002827924 +- 0.000001172 Vs BL = 0.05655847 +- 0.00002344 Tm Remeasure 17:44:34 Integrated Voltage Measurement... 17:44:34 Magnet Current for MAIN, Imag = 154.9747 A Vtrans = 1.549800 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:44:42 Integrated Voltage Measurement X0 = 0.0300000 m, Delta X = 0.0050000 m VT = 0.002821978 Vs 17:44:49 Integrated Voltage Measurement X0 = 0.0300000 m, Delta X = -0.0050000 m VT = -0.002838537 Vs 17:44:50 Magnet Current for MAIN, Imag = 154.9750 A Vtrans = 1.549803 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:44:58 Integrated Voltage Measurement X0 = 0.0300000 m, Delta X = 0.0050000 m VT = 0.002820260 Vs 17:45:05 Integrated Voltage Measurement X0 = 0.0300000 m, Delta X = -0.0050000 m VT = -0.002838048 Vs 17:45:06 Magnet Current for MAIN, Imag = 154.9747 A Vtrans = 1.549800 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:45:14 Integrated Voltage Measurement X0 = 0.0300000 m, Delta X = 0.0050000 m VT = 0.002820751 Vs 17:45:21 Integrated Voltage Measurement X0 = 0.0300000 m, Delta X = -0.0050000 m VT = -0.002834579 Vs 17:45:22 Magnet Current for MAIN, Imag = 154.9750 A Vtrans = 1.549803 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:45:30 Integrated Voltage Measurement X0 = 0.0300000 m, Delta X = 0.0050000 m VT = 0.002819257 Vs 17:45:37 Integrated Voltage Measurement X0 = 0.0300000 m, Delta X = -0.0050000 m VT = -0.002835507 Vs 17:45:37 Integrated Voltage Average and RMS of Measurements X0 = 0.0300000 m, Delta X = 0.0050000 m VT = 0.002828615 +- 0.000001163 Vs 17:45:37 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0300000 m, dx = 0.0050000 m, Nturns = 10.0 VT = 0.002828615 +- 0.000001163 Vs BL = 0.05657229 +- 0.00002326 Tm 17:45:37 Integrated Voltage Measurement... 17:45:38 Magnet Current for MAIN, Imag = 154.9751 A Vtrans = 1.549804 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:45:46 Integrated Voltage Measurement X0 = 0.0350000 m, Delta X = 0.0050000 m VT = 0.002673615 Vs 17:45:54 Integrated Voltage Measurement X0 = 0.0350000 m, Delta X = -0.0050000 m VT = -0.002679801 Vs 17:45:55 Magnet Current for MAIN, Imag = 154.9752 A Vtrans = 1.549805 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:46:02 Integrated Voltage Measurement X0 = 0.0350000 m, Delta X = 0.0050000 m VT = 0.002665100 Vs 17:46:10 Integrated Voltage Measurement X0 = 0.0350000 m, Delta X = -0.0050000 m VT = -0.002676126 Vs 17:46:11 Magnet Current for MAIN, Imag = 154.9750 A Vtrans = 1.549803 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:46:18 Integrated Voltage Measurement X0 = 0.0350000 m, Delta X = 0.0050000 m VT = 0.002664711 Vs 17:46:26 Integrated Voltage Measurement X0 = 0.0350000 m, Delta X = -0.0050000 m VT = -0.002685413 Vs 17:46:27 Magnet Current for MAIN, Imag = 154.9751 A Vtrans = 1.549804 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:46:34 Integrated Voltage Measurement X0 = 0.0350000 m, Delta X = 0.0050000 m VT = 0.002658940 Vs 17:46:42 Integrated Voltage Measurement X0 = 0.0350000 m, Delta X = -0.0050000 m VT = -0.002678120 Vs 17:46:42 Integrated Voltage Average and RMS of Measurements X0 = 0.0350000 m, Delta X = 0.0050000 m VT = 0.002672728 +- 0.000003293 Vs 17:46:42 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0350000 m, dx = 0.0050000 m, Nturns = 10.0 VT = 0.002672728 +- 0.000003293 Vs BL = 0.05345457 +- 0.00006587 Tm Remeasure 17:46:42 Integrated Voltage Measurement... 17:46:43 Magnet Current for MAIN, Imag = 154.9753 A Vtrans = 1.549807 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:46:50 Integrated Voltage Measurement X0 = 0.0350000 m, Delta X = 0.0050000 m VT = 0.002669014 Vs 17:46:58 Integrated Voltage Measurement X0 = 0.0350000 m, Delta X = -0.0050000 m VT = -0.002677758 Vs 17:46:59 Magnet Current for MAIN, Imag = 154.9753 A Vtrans = 1.549807 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:47:06 Integrated Voltage Measurement X0 = 0.0350000 m, Delta X = 0.0050000 m VT = 0.002663333 Vs 17:47:14 Integrated Voltage Measurement X0 = 0.0350000 m, Delta X = -0.0050000 m VT = -0.002669584 Vs 17:47:15 Magnet Current for MAIN, Imag = 154.9752 A Vtrans = 1.549805 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:47:22 Integrated Voltage Measurement X0 = 0.0350000 m, Delta X = 0.0050000 m VT = 0.002666547 Vs 17:47:30 Integrated Voltage Measurement X0 = 0.0350000 m, Delta X = -0.0050000 m VT = -0.002681807 Vs 17:47:30 Magnet Current for MAIN, Imag = 154.9753 A Vtrans = 1.549807 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:47:38 Integrated Voltage Measurement X0 = 0.0350000 m, Delta X = 0.0050000 m VT = 0.002660389 Vs 17:47:46 Integrated Voltage Measurement X0 = 0.0350000 m, Delta X = -0.0050000 m VT = -0.002681186 Vs 17:47:46 Integrated Voltage Average and RMS of Measurements X0 = 0.0350000 m, Delta X = 0.0050000 m VT = 0.002671202 +- 0.000003012 Vs 17:47:46 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0350000 m, dx = 0.0050000 m, Nturns = 10.0 VT = 0.002671202 +- 0.000003012 Vs BL = 0.05342405 +- 0.00006025 Tm Remeasure 17:47:46 Integrated Voltage Measurement... 17:47:47 Magnet Current for MAIN, Imag = 154.9755 A Vtrans = 1.549808 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:47:54 Integrated Voltage Measurement X0 = 0.0350000 m, Delta X = 0.0050000 m VT = 0.002667648 Vs 17:48:02 Integrated Voltage Measurement X0 = 0.0350000 m, Delta X = -0.0050000 m VT = -0.002674255 Vs 17:48:02 Magnet Current for MAIN, Imag = 154.9755 A Vtrans = 1.549808 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:48:10 Integrated Voltage Measurement X0 = 0.0350000 m, Delta X = 0.0050000 m VT = 0.002660127 Vs 17:48:18 Integrated Voltage Measurement X0 = 0.0350000 m, Delta X = -0.0050000 m VT = -0.002679350 Vs 17:48:18 Magnet Current for MAIN, Imag = 154.9752 A Vtrans = 1.549805 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:48:26 Integrated Voltage Measurement X0 = 0.0350000 m, Delta X = 0.0050000 m VT = 0.002667656 Vs 17:48:34 Integrated Voltage Measurement X0 = 0.0350000 m, Delta X = -0.0050000 m VT = -0.002683562 Vs 17:48:34 Magnet Current for MAIN, Imag = 154.9755 A Vtrans = 1.549808 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:48:42 Integrated Voltage Measurement X0 = 0.0350000 m, Delta X = 0.0050000 m VT = 0.002663806 Vs 17:48:49 Integrated Voltage Measurement X0 = 0.0350000 m, Delta X = -0.0050000 m VT = -0.002682205 Vs 17:48:49 Integrated Voltage Average and RMS of Measurements X0 = 0.0350000 m, Delta X = 0.0050000 m VT = 0.002672326 +- 0.000002226 Vs 17:48:50 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0350000 m, dx = 0.0050000 m, Nturns = 10.0 VT = 0.002672326 +- 0.000002226 Vs BL = 0.05344652 +- 0.00004453 Tm Remeasure 17:48:50 Integrated Voltage Measurement... 17:48:50 Magnet Current for MAIN, Imag = 154.9747 A Vtrans = 1.549800 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:48:58 Integrated Voltage Measurement X0 = 0.0350000 m, Delta X = 0.0050000 m VT = 0.002660904 Vs 17:49:05 Integrated Voltage Measurement X0 = 0.0350000 m, Delta X = -0.0050000 m VT = -0.002676841 Vs 17:49:06 Magnet Current for MAIN, Imag = 154.9750 A Vtrans = 1.549803 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:49:14 Integrated Voltage Measurement X0 = 0.0350000 m, Delta X = 0.0050000 m VT = 0.002665614 Vs 17:49:21 Integrated Voltage Measurement X0 = 0.0350000 m, Delta X = -0.0050000 m VT = -0.002679395 Vs 17:49:22 Magnet Current for MAIN, Imag = 154.9750 A Vtrans = 1.549803 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:49:30 Integrated Voltage Measurement X0 = 0.0350000 m, Delta X = 0.0050000 m VT = 0.002665984 Vs 17:49:37 Integrated Voltage Measurement X0 = 0.0350000 m, Delta X = -0.0050000 m VT = -0.002676548 Vs 17:49:38 Magnet Current for MAIN, Imag = 154.9748 A Vtrans = 1.549802 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:49:46 Integrated Voltage Measurement X0 = 0.0350000 m, Delta X = 0.0050000 m VT = 0.002667289 Vs 17:49:53 Integrated Voltage Measurement X0 = 0.0350000 m, Delta X = -0.0050000 m VT = -0.002680474 Vs 17:49:53 Integrated Voltage Average and RMS of Measurements X0 = 0.0350000 m, Delta X = 0.0050000 m VT = 0.002671631 +- 0.000001842 Vs 17:49:53 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0350000 m, dx = 0.0050000 m, Nturns = 10.0 VT = 0.002671631 +- 0.000001842 Vs BL = 0.05343263 +- 0.00003684 Tm Remeasure 17:49:53 Integrated Voltage Measurement... 17:49:54 Magnet Current for MAIN, Imag = 154.9751 A Vtrans = 1.549804 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:50:02 Integrated Voltage Measurement X0 = 0.0350000 m, Delta X = 0.0050000 m VT = 0.002659274 Vs 17:50:09 Integrated Voltage Measurement X0 = 0.0350000 m, Delta X = -0.0050000 m VT = -0.002674782 Vs 17:50:10 Magnet Current for MAIN, Imag = 154.9750 A Vtrans = 1.549803 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:50:18 Integrated Voltage Measurement X0 = 0.0350000 m, Delta X = 0.0050000 m VT = 0.002662771 Vs 17:50:25 Integrated Voltage Measurement X0 = 0.0350000 m, Delta X = -0.0050000 m VT = -0.002678745 Vs 17:50:26 Magnet Current for MAIN, Imag = 154.9747 A Vtrans = 1.549800 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:50:34 Integrated Voltage Measurement X0 = 0.0350000 m, Delta X = 0.0050000 m VT = 0.002664402 Vs 17:50:41 Integrated Voltage Measurement X0 = 0.0350000 m, Delta X = -0.0050000 m VT = -0.002679267 Vs 17:50:42 Magnet Current for MAIN, Imag = 154.9750 A Vtrans = 1.549803 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:50:49 Integrated Voltage Measurement X0 = 0.0350000 m, Delta X = 0.0050000 m VT = 0.002661328 Vs 17:50:57 Integrated Voltage Measurement X0 = 0.0350000 m, Delta X = -0.0050000 m VT = -0.002676786 Vs 17:50:57 Integrated Voltage Average and RMS of Measurements X0 = 0.0350000 m, Delta X = 0.0050000 m VT = 0.002669669 +- 0.000001818 Vs 17:50:57 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0350000 m, dx = 0.0050000 m, Nturns = 10.0 VT = 0.002669669 +- 0.000001818 Vs BL = 0.05339338 +- 0.00003637 Tm 17:51:01 Position moved: 0.000000 m 17:51:01 Magnet system MAIN current ramping to 155.0000 A... 17:51:12 Magnet Current for MAIN, Imag = 154.9752 A Vtrans = 1.549805 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:51:13 MAIN (V) Voltage Measurement, V = 6.113391 V 17:51:16 Temperature Measurement, Probe Number = 0, T = 26.203000 C 17:51:17 Temperature Measurement, Probe Number = 1, T = 24.279000 C 17:51:17 Temperature Measurement, Probe Number = 2, T = 28.191000 C 17:51:18 Temperature Measurement, Probe Number = 3, T = 30.377000 C 17:51:19 Temperature Measurement, Probe Number = 4, T = 28.705000 C 17:51:19 Temperature Measurement, Probe Number = 5, T = 24.501000 C 17:51:20 Sensor Voltage Measurement, Sensor Number = 0, V = 2.651551 V 17:51:20 Sensor Measurement, Sensor Number = 0, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 2.651551 (V, A, ohm, ...), Measured Value = 123.866332 (C, atm, Hz, ...) 17:51:20 Sensor Voltage Measurement, Sensor Number = 1, V = 1.158814 V 17:51:20 Sensor Measurement, Sensor Number = 1, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 1.158814 (V, A, ohm, ...), Measured Value = 11.911028 (C, atm, Hz, ...) 17:51:20 Magnet Test Parameter Measurements P0 = MAIN, Value = 154.98 P1 = MAIN (V), Value = 6.11 P2 = Yoke, Value = 26.20 P3 = Bottom Coil, Value = 24.28 P4 = Top Coil, Value = 28.19 P5 = Water In, Value = 30.38 P6 = Water Out, Value = 28.70 P7 = Ambient, Value = 24.50 P8 = Supply Pres(psi), Value = 123.87 P9 = Return Pres(psi), Value = 11.91 17:51:20 Integrated Voltage Measurement... 17:51:21 Magnet Current for MAIN, Imag = 154.9748 A Vtrans = 1.549802 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:51:30 Integrated Voltage Measurement X0 = -0.0350000 m, Delta X = 0.0050000 m VT = 0.002653912 Vs 17:51:37 Integrated Voltage Measurement X0 = -0.0350000 m, Delta X = -0.0050000 m VT = -0.002669566 Vs 17:51:38 Magnet Current for MAIN, Imag = 154.9748 A Vtrans = 1.549802 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:51:46 Integrated Voltage Measurement X0 = -0.0350000 m, Delta X = 0.0050000 m VT = 0.002644534 Vs 17:51:53 Integrated Voltage Measurement X0 = -0.0350000 m, Delta X = -0.0050000 m VT = -0.002674454 Vs 17:51:54 Magnet Current for MAIN, Imag = 154.9748 A Vtrans = 1.549802 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:52:02 Integrated Voltage Measurement X0 = -0.0350000 m, Delta X = 0.0050000 m VT = 0.002652353 Vs 17:52:09 Integrated Voltage Measurement X0 = -0.0350000 m, Delta X = -0.0050000 m VT = -0.002673232 Vs 17:52:10 Magnet Current for MAIN, Imag = 154.9757 A Vtrans = 1.549810 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:52:18 Integrated Voltage Measurement X0 = -0.0350000 m, Delta X = 0.0050000 m VT = 0.002661398 Vs 17:52:25 Integrated Voltage Measurement X0 = -0.0350000 m, Delta X = -0.0050000 m VT = -0.002666918 Vs 17:52:25 Integrated Voltage Average and RMS of Measurements X0 = -0.0350000 m, Delta X = 0.0050000 m VT = 0.002662046 +- 0.000001705 Vs 17:52:25 Integrated Field Strength, Average and RMS of Measurements X0 = -0.0350000 m, dx = 0.0050000 m, Nturns = 10.0 VT = 0.002662046 +- 0.000001705 Vs BL = 0.05324092 +- 0.00003409 Tm Remeasure 17:52:25 Integrated Voltage Measurement... 17:52:26 Magnet Current for MAIN, Imag = 154.9746 A Vtrans = 1.549799 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:52:33 Integrated Voltage Measurement X0 = -0.0350000 m, Delta X = 0.0050000 m VT = 0.002650715 Vs 17:52:41 Integrated Voltage Measurement X0 = -0.0350000 m, Delta X = -0.0050000 m VT = -0.002660797 Vs 17:52:42 Magnet Current for MAIN, Imag = 154.9751 A Vtrans = 1.549804 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:52:49 Integrated Voltage Measurement X0 = -0.0350000 m, Delta X = 0.0050000 m VT = 0.002654616 Vs 17:52:57 Integrated Voltage Measurement X0 = -0.0350000 m, Delta X = -0.0050000 m VT = -0.002664162 Vs 17:52:58 Magnet Current for MAIN, Imag = 154.9750 A Vtrans = 1.549803 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:53:05 Integrated Voltage Measurement X0 = -0.0350000 m, Delta X = 0.0050000 m VT = 0.002661700 Vs 17:53:13 Integrated Voltage Measurement X0 = -0.0350000 m, Delta X = -0.0050000 m VT = -0.002667012 Vs 17:53:14 Magnet Current for MAIN, Imag = 154.9753 A Vtrans = 1.549807 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:53:21 Integrated Voltage Measurement X0 = -0.0350000 m, Delta X = 0.0050000 m VT = 0.002654677 Vs 17:53:29 Integrated Voltage Measurement X0 = -0.0350000 m, Delta X = -0.0050000 m VT = -0.002660656 Vs 17:53:29 Integrated Voltage Average and RMS of Measurements X0 = -0.0350000 m, Delta X = 0.0050000 m VT = 0.002659292 +- 0.000003194 Vs 17:53:29 Integrated Field Strength, Average and RMS of Measurements X0 = -0.0350000 m, dx = 0.0050000 m, Nturns = 10.0 VT = 0.002659292 +- 0.000003194 Vs BL = 0.05318584 +- 0.00006388 Tm Remeasure 17:53:29 Integrated Voltage Measurement... 17:53:29 Magnet Current for MAIN, Imag = 154.9750 A Vtrans = 1.549803 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:53:37 Integrated Voltage Measurement X0 = -0.0350000 m, Delta X = 0.0050000 m VT = 0.002659536 Vs 17:53:45 Integrated Voltage Measurement X0 = -0.0350000 m, Delta X = -0.0050000 m VT = -0.002665285 Vs 17:53:45 Magnet Current for MAIN, Imag = 154.9747 A Vtrans = 1.549800 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:53:53 Integrated Voltage Measurement X0 = -0.0350000 m, Delta X = 0.0050000 m VT = 0.002651705 Vs 17:54:01 Integrated Voltage Measurement X0 = -0.0350000 m, Delta X = -0.0050000 m VT = -0.002673867 Vs 17:54:01 Magnet Current for MAIN, Imag = 154.9751 A Vtrans = 1.549804 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:54:09 Integrated Voltage Measurement X0 = -0.0350000 m, Delta X = 0.0050000 m VT = 0.002650258 Vs 17:54:17 Integrated Voltage Measurement X0 = -0.0350000 m, Delta X = -0.0050000 m VT = -0.002670523 Vs 17:54:17 Magnet Current for MAIN, Imag = 154.9747 A Vtrans = 1.549800 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:54:25 Integrated Voltage Measurement X0 = -0.0350000 m, Delta X = 0.0050000 m VT = 0.002652182 Vs 17:54:33 Integrated Voltage Measurement X0 = -0.0350000 m, Delta X = -0.0050000 m VT = -0.002670879 Vs 17:54:33 Integrated Voltage Average and RMS of Measurements X0 = -0.0350000 m, Delta X = 0.0050000 m VT = 0.002661779 +- 0.000000922 Vs 17:54:33 Integrated Field Strength, Average and RMS of Measurements X0 = -0.0350000 m, dx = 0.0050000 m, Nturns = 10.0 VT = 0.002661779 +- 0.000000922 Vs BL = 0.05323559 +- 0.00001845 Tm Remeasure 17:54:33 Integrated Voltage Measurement... 17:54:33 Magnet Current for MAIN, Imag = 154.9755 A Vtrans = 1.549808 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:54:41 Integrated Voltage Measurement X0 = -0.0350000 m, Delta X = 0.0050000 m VT = 0.002657115 Vs 17:54:49 Integrated Voltage Measurement X0 = -0.0350000 m, Delta X = -0.0050000 m VT = -0.002669201 Vs 17:54:49 Magnet Current for MAIN, Imag = 154.9753 A Vtrans = 1.549807 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:54:57 Integrated Voltage Measurement X0 = -0.0350000 m, Delta X = 0.0050000 m VT = 0.002651041 Vs 17:55:05 Integrated Voltage Measurement X0 = -0.0350000 m, Delta X = -0.0050000 m VT = -0.002680434 Vs 17:55:05 Magnet Current for MAIN, Imag = 154.9751 A Vtrans = 1.549804 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:55:13 Integrated Voltage Measurement X0 = -0.0350000 m, Delta X = 0.0050000 m VT = 0.002653293 Vs 17:55:20 Integrated Voltage Measurement X0 = -0.0350000 m, Delta X = -0.0050000 m VT = -0.002661190 Vs 17:55:21 Magnet Current for MAIN, Imag = 154.9751 A Vtrans = 1.549804 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:55:29 Integrated Voltage Measurement X0 = -0.0350000 m, Delta X = 0.0050000 m VT = 0.002648153 Vs 17:55:36 Integrated Voltage Measurement X0 = -0.0350000 m, Delta X = -0.0050000 m VT = -0.002665991 Vs 17:55:36 Integrated Voltage Average and RMS of Measurements X0 = -0.0350000 m, Delta X = 0.0050000 m VT = 0.002660802 +- 0.000003758 Vs 17:55:36 Integrated Field Strength, Average and RMS of Measurements X0 = -0.0350000 m, dx = 0.0050000 m, Nturns = 10.0 VT = 0.002660802 +- 0.000003758 Vs BL = 0.05321604 +- 0.00007517 Tm Remeasure 17:55:36 Integrated Voltage Measurement... 17:55:37 Magnet Current for MAIN, Imag = 154.9748 A Vtrans = 1.549802 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:55:45 Integrated Voltage Measurement X0 = -0.0350000 m, Delta X = 0.0050000 m VT = 0.002659324 Vs 17:55:52 Integrated Voltage Measurement X0 = -0.0350000 m, Delta X = -0.0050000 m VT = -0.002667981 Vs 17:55:53 Magnet Current for MAIN, Imag = 154.9755 A Vtrans = 1.549808 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:56:00 Integrated Voltage Measurement X0 = -0.0350000 m, Delta X = 0.0050000 m VT = 0.002656183 Vs 17:56:08 Integrated Voltage Measurement X0 = -0.0350000 m, Delta X = -0.0050000 m VT = -0.002672462 Vs 17:56:09 Magnet Current for MAIN, Imag = 154.9748 A Vtrans = 1.549802 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:56:17 Integrated Voltage Measurement X0 = -0.0350000 m, Delta X = 0.0050000 m VT = 0.002651888 Vs 17:56:24 Integrated Voltage Measurement X0 = -0.0350000 m, Delta X = -0.0050000 m VT = -0.002669018 Vs 17:56:25 Magnet Current for MAIN, Imag = 154.9748 A Vtrans = 1.549802 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:56:33 Integrated Voltage Measurement X0 = -0.0350000 m, Delta X = 0.0050000 m VT = 0.002650035 Vs 17:56:40 Integrated Voltage Measurement X0 = -0.0350000 m, Delta X = -0.0050000 m VT = -0.002660447 Vs 17:56:40 Integrated Voltage Average and RMS of Measurements X0 = -0.0350000 m, Delta X = 0.0050000 m VT = 0.002660917 +- 0.000003589 Vs 17:56:40 Integrated Field Strength, Average and RMS of Measurements X0 = -0.0350000 m, dx = 0.0050000 m, Nturns = 10.0 VT = 0.002660917 +- 0.000003589 Vs BL = 0.05321835 +- 0.00007177 Tm 17:56:40 Integrated Voltage Measurement... 17:56:41 Magnet Current for MAIN, Imag = 154.9750 A Vtrans = 1.549803 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:56:50 Integrated Voltage Measurement X0 = -0.0300000 m, Delta X = 0.0050000 m VT = 0.002823919 Vs 17:56:57 Integrated Voltage Measurement X0 = -0.0300000 m, Delta X = -0.0050000 m VT = -0.002826293 Vs 17:56:58 Magnet Current for MAIN, Imag = 154.9748 A Vtrans = 1.549802 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:57:06 Integrated Voltage Measurement X0 = -0.0300000 m, Delta X = 0.0050000 m VT = 0.002818822 Vs 17:57:13 Integrated Voltage Measurement X0 = -0.0300000 m, Delta X = -0.0050000 m VT = -0.002837218 Vs 17:57:14 Magnet Current for MAIN, Imag = 154.9748 A Vtrans = 1.549802 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:57:22 Integrated Voltage Measurement X0 = -0.0300000 m, Delta X = 0.0050000 m VT = 0.002813460 Vs 17:57:29 Integrated Voltage Measurement X0 = -0.0300000 m, Delta X = -0.0050000 m VT = -0.002829772 Vs 17:57:30 Magnet Current for MAIN, Imag = 154.9750 A Vtrans = 1.549803 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:57:37 Integrated Voltage Measurement X0 = -0.0300000 m, Delta X = 0.0050000 m VT = 0.002818058 Vs 17:57:45 Integrated Voltage Measurement X0 = -0.0300000 m, Delta X = -0.0050000 m VT = -0.002833368 Vs 17:57:45 Integrated Voltage Average and RMS of Measurements X0 = -0.0300000 m, Delta X = 0.0050000 m VT = 0.002825114 +- 0.000002294 Vs 17:57:45 Integrated Field Strength, Average and RMS of Measurements X0 = -0.0300000 m, dx = 0.0050000 m, Nturns = 10.0 VT = 0.002825114 +- 0.000002294 Vs BL = 0.05650228 +- 0.00004587 Tm Remeasure 17:57:45 Integrated Voltage Measurement... 17:57:46 Magnet Current for MAIN, Imag = 154.9752 A Vtrans = 1.549805 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:57:53 Integrated Voltage Measurement X0 = -0.0300000 m, Delta X = 0.0050000 m VT = 0.002821715 Vs 17:58:01 Integrated Voltage Measurement X0 = -0.0300000 m, Delta X = -0.0050000 m VT = -0.002833046 Vs 17:58:02 Magnet Current for MAIN, Imag = 154.9752 A Vtrans = 1.549805 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:58:09 Integrated Voltage Measurement X0 = -0.0300000 m, Delta X = 0.0050000 m VT = 0.002812771 Vs 17:58:17 Integrated Voltage Measurement X0 = -0.0300000 m, Delta X = -0.0050000 m VT = -0.002836508 Vs 17:58:18 Magnet Current for MAIN, Imag = 154.9751 A Vtrans = 1.549804 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:58:25 Integrated Voltage Measurement X0 = -0.0300000 m, Delta X = 0.0050000 m VT = 0.002821329 Vs 17:58:33 Integrated Voltage Measurement X0 = -0.0300000 m, Delta X = -0.0050000 m VT = -0.002833902 Vs 17:58:34 Magnet Current for MAIN, Imag = 154.9751 A Vtrans = 1.549804 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:58:41 Integrated Voltage Measurement X0 = -0.0300000 m, Delta X = 0.0050000 m VT = 0.002814561 Vs 17:58:49 Integrated Voltage Measurement X0 = -0.0300000 m, Delta X = -0.0050000 m VT = -0.002828852 Vs 17:58:49 Integrated Voltage Average and RMS of Measurements X0 = -0.0300000 m, Delta X = 0.0050000 m VT = 0.002825336 +- 0.000002400 Vs 17:58:49 Integrated Field Strength, Average and RMS of Measurements X0 = -0.0300000 m, dx = 0.0050000 m, Nturns = 10.0 VT = 0.002825336 +- 0.000002400 Vs BL = 0.05650671 +- 0.00004799 Tm Remeasure 17:58:49 Integrated Voltage Measurement... 17:58:50 Magnet Current for MAIN, Imag = 154.9750 A Vtrans = 1.549803 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:58:57 Integrated Voltage Measurement X0 = -0.0300000 m, Delta X = 0.0050000 m VT = 0.002814472 Vs 17:59:05 Integrated Voltage Measurement X0 = -0.0300000 m, Delta X = -0.0050000 m VT = -0.002839901 Vs 17:59:06 Magnet Current for MAIN, Imag = 154.9747 A Vtrans = 1.549800 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:59:13 Integrated Voltage Measurement X0 = -0.0300000 m, Delta X = 0.0050000 m VT = 0.002816799 Vs 17:59:21 Integrated Voltage Measurement X0 = -0.0300000 m, Delta X = -0.0050000 m VT = -0.002831904 Vs 17:59:21 Magnet Current for MAIN, Imag = 154.9746 A Vtrans = 1.549799 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:59:29 Integrated Voltage Measurement X0 = -0.0300000 m, Delta X = 0.0050000 m VT = 0.002811302 Vs 17:59:36 Integrated Voltage Measurement X0 = -0.0300000 m, Delta X = -0.0050000 m VT = -0.002833730 Vs 17:59:37 Magnet Current for MAIN, Imag = 154.9753 A Vtrans = 1.549807 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 17:59:45 Integrated Voltage Measurement X0 = -0.0300000 m, Delta X = 0.0050000 m VT = 0.002814686 Vs 17:59:52 Integrated Voltage Measurement X0 = -0.0300000 m, Delta X = -0.0050000 m VT = -0.002834896 Vs 17:59:52 Integrated Voltage Average and RMS of Measurements X0 = -0.0300000 m, Delta X = 0.0050000 m VT = 0.002824711 +- 0.000001664 Vs 17:59:52 Integrated Field Strength, Average and RMS of Measurements X0 = -0.0300000 m, dx = 0.0050000 m, Nturns = 10.0 VT = 0.002824711 +- 0.000001664 Vs BL = 0.05649423 +- 0.00003329 Tm Remeasure 17:59:52 Integrated Voltage Measurement... 17:59:53 Magnet Current for MAIN, Imag = 154.9750 A Vtrans = 1.549803 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:00:01 Integrated Voltage Measurement X0 = -0.0300000 m, Delta X = 0.0050000 m VT = 0.002817289 Vs 18:00:08 Integrated Voltage Measurement X0 = -0.0300000 m, Delta X = -0.0050000 m VT = -0.002825611 Vs 18:00:09 Magnet Current for MAIN, Imag = 154.9747 A Vtrans = 1.549800 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:00:17 Integrated Voltage Measurement X0 = -0.0300000 m, Delta X = 0.0050000 m VT = 0.002818534 Vs 18:00:24 Integrated Voltage Measurement X0 = -0.0300000 m, Delta X = -0.0050000 m VT = -0.002831820 Vs 18:00:25 Magnet Current for MAIN, Imag = 154.9750 A Vtrans = 1.549803 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:00:33 Integrated Voltage Measurement X0 = -0.0300000 m, Delta X = 0.0050000 m VT = 0.002811361 Vs 18:00:40 Integrated Voltage Measurement X0 = -0.0300000 m, Delta X = -0.0050000 m VT = -0.002827739 Vs 18:00:41 Magnet Current for MAIN, Imag = 154.9750 A Vtrans = 1.549803 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:00:49 Integrated Voltage Measurement X0 = -0.0300000 m, Delta X = 0.0050000 m VT = 0.002814492 Vs 18:00:56 Integrated Voltage Measurement X0 = -0.0300000 m, Delta X = -0.0050000 m VT = -0.002823190 Vs 18:00:56 Integrated Voltage Average and RMS of Measurements X0 = -0.0300000 m, Delta X = 0.0050000 m VT = 0.002821254 +- 0.000002457 Vs 18:00:56 Integrated Field Strength, Average and RMS of Measurements X0 = -0.0300000 m, dx = 0.0050000 m, Nturns = 10.0 VT = 0.002821254 +- 0.000002457 Vs BL = 0.05642509 +- 0.00004915 Tm Remeasure 18:00:56 Integrated Voltage Measurement... 18:00:57 Magnet Current for MAIN, Imag = 154.9748 A Vtrans = 1.549802 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:01:05 Integrated Voltage Measurement X0 = -0.0300000 m, Delta X = 0.0050000 m VT = 0.002817040 Vs 18:01:12 Integrated Voltage Measurement X0 = -0.0300000 m, Delta X = -0.0050000 m VT = -0.002838347 Vs 18:01:13 Magnet Current for MAIN, Imag = 154.9746 A Vtrans = 1.549799 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:01:20 Integrated Voltage Measurement X0 = -0.0300000 m, Delta X = 0.0050000 m VT = 0.002819944 Vs 18:01:28 Integrated Voltage Measurement X0 = -0.0300000 m, Delta X = -0.0050000 m VT = -0.002837109 Vs 18:01:29 Magnet Current for MAIN, Imag = 154.9750 A Vtrans = 1.549803 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:01:36 Integrated Voltage Measurement X0 = -0.0300000 m, Delta X = 0.0050000 m VT = 0.002815044 Vs 18:01:44 Integrated Voltage Measurement X0 = -0.0300000 m, Delta X = -0.0050000 m VT = -0.002832604 Vs 18:01:45 Magnet Current for MAIN, Imag = 154.9748 A Vtrans = 1.549802 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:01:52 Integrated Voltage Measurement X0 = -0.0300000 m, Delta X = 0.0050000 m VT = 0.002812111 Vs 18:02:00 Integrated Voltage Measurement X0 = -0.0300000 m, Delta X = -0.0050000 m VT = -0.002837222 Vs 18:02:00 Integrated Voltage Average and RMS of Measurements X0 = -0.0300000 m, Delta X = 0.0050000 m VT = 0.002826178 +- 0.000001977 Vs 18:02:00 Integrated Field Strength, Average and RMS of Measurements X0 = -0.0300000 m, dx = 0.0050000 m, Nturns = 10.0 VT = 0.002826178 +- 0.000001977 Vs BL = 0.05652356 +- 0.00003955 Tm 18:02:00 Integrated Voltage Measurement... 18:02:01 Magnet Current for MAIN, Imag = 154.9747 A Vtrans = 1.549800 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:02:10 Integrated Voltage Measurement X0 = -0.0250000 m, Delta X = 0.0050000 m VT = 0.002894897 Vs 18:02:17 Integrated Voltage Measurement X0 = -0.0250000 m, Delta X = -0.0050000 m VT = -0.002915939 Vs 18:02:18 Magnet Current for MAIN, Imag = 154.9752 A Vtrans = 1.549805 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:02:26 Integrated Voltage Measurement X0 = -0.0250000 m, Delta X = 0.0050000 m VT = 0.002899220 Vs 18:02:33 Integrated Voltage Measurement X0 = -0.0250000 m, Delta X = -0.0050000 m VT = -0.002915414 Vs 18:02:34 Magnet Current for MAIN, Imag = 154.9745 A Vtrans = 1.549798 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:02:42 Integrated Voltage Measurement X0 = -0.0250000 m, Delta X = 0.0050000 m VT = 0.002898194 Vs 18:02:49 Integrated Voltage Measurement X0 = -0.0250000 m, Delta X = -0.0050000 m VT = -0.002914418 Vs 18:02:50 Magnet Current for MAIN, Imag = 154.9753 A Vtrans = 1.549807 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:02:57 Integrated Voltage Measurement X0 = -0.0250000 m, Delta X = 0.0050000 m VT = 0.002903890 Vs 18:03:05 Integrated Voltage Measurement X0 = -0.0250000 m, Delta X = -0.0050000 m VT = -0.002916364 Vs 18:03:05 Integrated Voltage Average and RMS of Measurements X0 = -0.0250000 m, Delta X = 0.0050000 m VT = 0.002907292 +- 0.000001769 Vs 18:03:05 Integrated Field Strength, Average and RMS of Measurements X0 = -0.0250000 m, dx = 0.0050000 m, Nturns = 10.0 VT = 0.002907292 +- 0.000001769 Vs BL = 0.05814584 +- 0.00003538 Tm Remeasure 18:03:05 Integrated Voltage Measurement... 18:03:06 Magnet Current for MAIN, Imag = 154.9748 A Vtrans = 1.549802 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:03:14 Integrated Voltage Measurement X0 = -0.0250000 m, Delta X = 0.0050000 m VT = 0.002898298 Vs 18:03:21 Integrated Voltage Measurement X0 = -0.0250000 m, Delta X = -0.0050000 m VT = -0.002915574 Vs 18:03:22 Magnet Current for MAIN, Imag = 154.9746 A Vtrans = 1.549799 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:03:29 Integrated Voltage Measurement X0 = -0.0250000 m, Delta X = 0.0050000 m VT = 0.002899749 Vs 18:03:37 Integrated Voltage Measurement X0 = -0.0250000 m, Delta X = -0.0050000 m VT = -0.002911681 Vs 18:03:38 Magnet Current for MAIN, Imag = 154.9757 A Vtrans = 1.549810 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:03:45 Integrated Voltage Measurement X0 = -0.0250000 m, Delta X = 0.0050000 m VT = 0.002899230 Vs 18:03:53 Integrated Voltage Measurement X0 = -0.0250000 m, Delta X = -0.0050000 m VT = -0.002915842 Vs 18:03:53 Magnet Current for MAIN, Imag = 154.9752 A Vtrans = 1.549805 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:04:01 Integrated Voltage Measurement X0 = -0.0250000 m, Delta X = 0.0050000 m VT = 0.002897465 Vs 18:04:09 Integrated Voltage Measurement X0 = -0.0250000 m, Delta X = -0.0050000 m VT = -0.002922779 Vs 18:04:09 Integrated Voltage Average and RMS of Measurements X0 = -0.0250000 m, Delta X = 0.0050000 m VT = 0.002907577 +- 0.000001609 Vs 18:04:09 Integrated Field Strength, Average and RMS of Measurements X0 = -0.0250000 m, dx = 0.0050000 m, Nturns = 10.0 VT = 0.002907577 +- 0.000001609 Vs BL = 0.05815154 +- 0.00003218 Tm Remeasure 18:04:09 Integrated Voltage Measurement... 18:04:09 Magnet Current for MAIN, Imag = 154.9753 A Vtrans = 1.549807 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:04:17 Integrated Voltage Measurement X0 = -0.0250000 m, Delta X = 0.0050000 m VT = 0.002902467 Vs 18:04:24 Integrated Voltage Measurement X0 = -0.0250000 m, Delta X = -0.0050000 m VT = -0.002912301 Vs 18:04:25 Magnet Current for MAIN, Imag = 154.9748 A Vtrans = 1.549802 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:04:33 Integrated Voltage Measurement X0 = -0.0250000 m, Delta X = 0.0050000 m VT = 0.002897427 Vs 18:04:40 Integrated Voltage Measurement X0 = -0.0250000 m, Delta X = -0.0050000 m VT = -0.002916768 Vs 18:04:41 Magnet Current for MAIN, Imag = 154.9752 A Vtrans = 1.549805 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:04:49 Integrated Voltage Measurement X0 = -0.0250000 m, Delta X = 0.0050000 m VT = 0.002899905 Vs 18:04:56 Integrated Voltage Measurement X0 = -0.0250000 m, Delta X = -0.0050000 m VT = -0.002915999 Vs 18:04:57 Magnet Current for MAIN, Imag = 154.9756 A Vtrans = 1.549809 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:05:05 Integrated Voltage Measurement X0 = -0.0250000 m, Delta X = 0.0050000 m VT = 0.002899683 Vs 18:05:12 Integrated Voltage Measurement X0 = -0.0250000 m, Delta X = -0.0050000 m VT = -0.002914968 Vs 18:05:12 Integrated Voltage Average and RMS of Measurements X0 = -0.0250000 m, Delta X = 0.0050000 m VT = 0.002907440 +- 0.000000315 Vs 18:05:12 Integrated Field Strength, Average and RMS of Measurements X0 = -0.0250000 m, dx = 0.0050000 m, Nturns = 10.0 VT = 0.002907440 +- 0.000000315 Vs BL = 0.05814880 +- 0.00000629 Tm 18:05:12 Integrated Voltage Measurement... 18:05:13 Magnet Current for MAIN, Imag = 154.9755 A Vtrans = 1.549808 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:05:22 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = 0.0050000 m VT = 0.002936597 Vs 18:05:29 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = -0.0050000 m VT = -0.002955431 Vs 18:05:30 Magnet Current for MAIN, Imag = 154.9755 A Vtrans = 1.549808 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:05:38 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = 0.0050000 m VT = 0.002935453 Vs 18:05:45 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = -0.0050000 m VT = -0.002954144 Vs 18:05:46 Magnet Current for MAIN, Imag = 154.9752 A Vtrans = 1.549805 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:05:54 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = 0.0050000 m VT = 0.002930437 Vs 18:06:01 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = -0.0050000 m VT = -0.002955265 Vs 18:06:02 Magnet Current for MAIN, Imag = 154.9750 A Vtrans = 1.549803 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:06:10 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = 0.0050000 m VT = 0.002948592 Vs 18:06:17 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = -0.0050000 m VT = -0.002953021 Vs 18:06:17 Integrated Voltage Average and RMS of Measurements X0 = -0.0200000 m, Delta X = 0.0050000 m VT = 0.002946117 +- 0.000002933 Vs 18:06:17 Integrated Field Strength, Average and RMS of Measurements X0 = -0.0200000 m, dx = 0.0050000 m, Nturns = 10.0 VT = 0.002946117 +- 0.000002933 Vs BL = 0.05892235 +- 0.00005866 Tm Remeasure 18:06:17 Integrated Voltage Measurement... 18:06:18 Magnet Current for MAIN, Imag = 154.9751 A Vtrans = 1.549804 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:06:25 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = 0.0050000 m VT = 0.002936928 Vs 18:06:33 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = -0.0050000 m VT = -0.002958303 Vs 18:06:34 Magnet Current for MAIN, Imag = 154.9753 A Vtrans = 1.549807 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:06:41 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = 0.0050000 m VT = 0.002948757 Vs 18:06:49 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = -0.0050000 m VT = -0.002955783 Vs 18:06:50 Magnet Current for MAIN, Imag = 154.9753 A Vtrans = 1.549807 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:06:57 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = 0.0050000 m VT = 0.002938995 Vs 18:07:05 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = -0.0050000 m VT = -0.002954327 Vs 18:07:06 Magnet Current for MAIN, Imag = 154.9748 A Vtrans = 1.549802 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:07:13 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = 0.0050000 m VT = 0.002933614 Vs 18:07:21 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = -0.0050000 m VT = -0.002953532 Vs 18:07:21 Integrated Voltage Average and RMS of Measurements X0 = -0.0200000 m, Delta X = 0.0050000 m VT = 0.002947530 +- 0.000003118 Vs 18:07:21 Integrated Field Strength, Average and RMS of Measurements X0 = -0.0200000 m, dx = 0.0050000 m, Nturns = 10.0 VT = 0.002947530 +- 0.000003118 Vs BL = 0.05895060 +- 0.00006236 Tm Remeasure 18:07:21 Integrated Voltage Measurement... 18:07:22 Magnet Current for MAIN, Imag = 154.9752 A Vtrans = 1.549805 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:07:29 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = 0.0050000 m VT = 0.002940151 Vs 18:07:37 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = -0.0050000 m VT = -0.002956405 Vs 18:07:38 Magnet Current for MAIN, Imag = 154.9751 A Vtrans = 1.549804 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:07:45 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = 0.0050000 m VT = 0.002942723 Vs 18:07:53 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = -0.0050000 m VT = -0.002952194 Vs 18:07:54 Magnet Current for MAIN, Imag = 154.9750 A Vtrans = 1.549803 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:08:01 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = 0.0050000 m VT = 0.002935093 Vs 18:08:09 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = -0.0050000 m VT = -0.002955044 Vs 18:08:10 Magnet Current for MAIN, Imag = 154.9751 A Vtrans = 1.549804 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:08:17 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = 0.0050000 m VT = 0.002936772 Vs 18:08:25 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = -0.0050000 m VT = -0.002954908 Vs 18:08:25 Integrated Voltage Average and RMS of Measurements X0 = -0.0200000 m, Delta X = 0.0050000 m VT = 0.002946661 +- 0.000001271 Vs 18:08:25 Integrated Field Strength, Average and RMS of Measurements X0 = -0.0200000 m, dx = 0.0050000 m, Nturns = 10.0 VT = 0.002946661 +- 0.000001271 Vs BL = 0.05893322 +- 0.00002542 Tm Remeasure 18:08:25 Integrated Voltage Measurement... 18:08:26 Magnet Current for MAIN, Imag = 154.9751 A Vtrans = 1.549804 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:08:33 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = 0.0050000 m VT = 0.002932216 Vs 18:08:41 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = -0.0050000 m VT = -0.002952132 Vs 18:08:42 Magnet Current for MAIN, Imag = 154.9751 A Vtrans = 1.549804 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:08:49 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = 0.0050000 m VT = 0.002937533 Vs 18:08:57 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = -0.0050000 m VT = -0.002950778 Vs 18:08:58 Magnet Current for MAIN, Imag = 154.9745 A Vtrans = 1.549798 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:09:05 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = 0.0050000 m VT = 0.002947308 Vs 18:09:13 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = -0.0050000 m VT = -0.002953522 Vs 18:09:13 Magnet Current for MAIN, Imag = 154.9751 A Vtrans = 1.549804 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:09:21 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = 0.0050000 m VT = 0.002936857 Vs 18:09:29 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = -0.0050000 m VT = -0.002954635 Vs 18:09:29 Integrated Voltage Average and RMS of Measurements X0 = -0.0200000 m, Delta X = 0.0050000 m VT = 0.002945623 +- 0.000003042 Vs 18:09:29 Integrated Field Strength, Average and RMS of Measurements X0 = -0.0200000 m, dx = 0.0050000 m, Nturns = 10.0 VT = 0.002945623 +- 0.000003042 Vs BL = 0.05891245 +- 0.00006085 Tm Remeasure 18:09:29 Integrated Voltage Measurement... 18:09:29 Magnet Current for MAIN, Imag = 154.9750 A Vtrans = 1.549803 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:09:37 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = 0.0050000 m VT = 0.002939260 Vs 18:09:45 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = -0.0050000 m VT = -0.002951323 Vs 18:09:45 Magnet Current for MAIN, Imag = 154.9751 A Vtrans = 1.549804 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:09:53 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = 0.0050000 m VT = 0.002934374 Vs 18:10:01 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = -0.0050000 m VT = -0.002954276 Vs 18:10:01 Magnet Current for MAIN, Imag = 154.9753 A Vtrans = 1.549807 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:10:09 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = 0.0050000 m VT = 0.002936001 Vs 18:10:16 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = -0.0050000 m VT = -0.002953887 Vs 18:10:17 Magnet Current for MAIN, Imag = 154.9751 A Vtrans = 1.549804 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:10:25 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = 0.0050000 m VT = 0.002941989 Vs 18:10:32 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = -0.0050000 m VT = -0.002952190 Vs 18:10:32 Integrated Voltage Average and RMS of Measurements X0 = -0.0200000 m, Delta X = 0.0050000 m VT = 0.002945412 +- 0.000001028 Vs 18:10:32 Integrated Field Strength, Average and RMS of Measurements X0 = -0.0200000 m, dx = 0.0050000 m, Nturns = 10.0 VT = 0.002945412 +- 0.000001028 Vs BL = 0.05890825 +- 0.00002057 Tm 18:10:32 Integrated Voltage Measurement... 18:10:33 Magnet Current for MAIN, Imag = 154.9752 A Vtrans = 1.549805 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:10:42 Integrated Voltage Measurement X0 = -0.0150000 m, Delta X = 0.0050000 m VT = 0.002952645 Vs 18:10:49 Integrated Voltage Measurement X0 = -0.0150000 m, Delta X = -0.0050000 m VT = -0.002962236 Vs 18:10:50 Magnet Current for MAIN, Imag = 154.9750 A Vtrans = 1.549803 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:10:58 Integrated Voltage Measurement X0 = -0.0150000 m, Delta X = 0.0050000 m VT = 0.002954072 Vs 18:11:05 Integrated Voltage Measurement X0 = -0.0150000 m, Delta X = -0.0050000 m VT = -0.002966794 Vs 18:11:06 Magnet Current for MAIN, Imag = 154.9752 A Vtrans = 1.549805 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:11:13 Integrated Voltage Measurement X0 = -0.0150000 m, Delta X = 0.0050000 m VT = 0.002954618 Vs 18:11:21 Integrated Voltage Measurement X0 = -0.0150000 m, Delta X = -0.0050000 m VT = -0.002973387 Vs 18:11:22 Magnet Current for MAIN, Imag = 154.9750 A Vtrans = 1.549803 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:11:29 Integrated Voltage Measurement X0 = -0.0150000 m, Delta X = 0.0050000 m VT = 0.002953419 Vs 18:11:37 Integrated Voltage Measurement X0 = -0.0150000 m, Delta X = -0.0050000 m VT = -0.002964099 Vs 18:11:37 Integrated Voltage Average and RMS of Measurements X0 = -0.0150000 m, Delta X = 0.0050000 m VT = 0.002960159 +- 0.000002459 Vs 18:11:37 Integrated Field Strength, Average and RMS of Measurements X0 = -0.0150000 m, dx = 0.0050000 m, Nturns = 10.0 VT = 0.002960159 +- 0.000002459 Vs BL = 0.05920317 +- 0.00004919 Tm Remeasure 18:11:37 Integrated Voltage Measurement... 18:11:38 Magnet Current for MAIN, Imag = 154.9752 A Vtrans = 1.549805 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:11:45 Integrated Voltage Measurement X0 = -0.0150000 m, Delta X = 0.0050000 m VT = 0.002953186 Vs 18:11:53 Integrated Voltage Measurement X0 = -0.0150000 m, Delta X = -0.0050000 m VT = -0.002960874 Vs 18:11:54 Magnet Current for MAIN, Imag = 154.9751 A Vtrans = 1.549804 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:12:01 Integrated Voltage Measurement X0 = -0.0150000 m, Delta X = 0.0050000 m VT = 0.002954138 Vs 18:12:09 Integrated Voltage Measurement X0 = -0.0150000 m, Delta X = -0.0050000 m VT = -0.002972119 Vs 18:12:10 Magnet Current for MAIN, Imag = 154.9748 A Vtrans = 1.549802 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:12:17 Integrated Voltage Measurement X0 = -0.0150000 m, Delta X = 0.0050000 m VT = 0.002951613 Vs 18:12:25 Integrated Voltage Measurement X0 = -0.0150000 m, Delta X = -0.0050000 m VT = -0.002970985 Vs 18:12:25 Magnet Current for MAIN, Imag = 154.9750 A Vtrans = 1.549803 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:12:33 Integrated Voltage Measurement X0 = -0.0150000 m, Delta X = 0.0050000 m VT = 0.002955085 Vs 18:12:41 Integrated Voltage Measurement X0 = -0.0150000 m, Delta X = -0.0050000 m VT = -0.002961722 Vs 18:12:41 Integrated Voltage Average and RMS of Measurements X0 = -0.0150000 m, Delta X = 0.0050000 m VT = 0.002959965 +- 0.000002390 Vs 18:12:41 Integrated Field Strength, Average and RMS of Measurements X0 = -0.0150000 m, dx = 0.0050000 m, Nturns = 10.0 VT = 0.002959965 +- 0.000002390 Vs BL = 0.05919930 +- 0.00004779 Tm Remeasure 18:12:41 Integrated Voltage Measurement... 18:12:41 Magnet Current for MAIN, Imag = 154.9748 A Vtrans = 1.549802 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:12:49 Integrated Voltage Measurement X0 = -0.0150000 m, Delta X = 0.0050000 m VT = 0.002954718 Vs 18:12:57 Integrated Voltage Measurement X0 = -0.0150000 m, Delta X = -0.0050000 m VT = -0.002967557 Vs 18:12:57 Magnet Current for MAIN, Imag = 154.9750 A Vtrans = 1.549803 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:13:05 Integrated Voltage Measurement X0 = -0.0150000 m, Delta X = 0.0050000 m VT = 0.002949867 Vs 18:13:12 Integrated Voltage Measurement X0 = -0.0150000 m, Delta X = -0.0050000 m VT = -0.002970817 Vs 18:13:13 Magnet Current for MAIN, Imag = 154.9750 A Vtrans = 1.549803 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:13:21 Integrated Voltage Measurement X0 = -0.0150000 m, Delta X = 0.0050000 m VT = 0.002952103 Vs 18:13:28 Integrated Voltage Measurement X0 = -0.0150000 m, Delta X = -0.0050000 m VT = -0.002964473 Vs 18:13:29 Magnet Current for MAIN, Imag = 154.9746 A Vtrans = 1.549799 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:13:37 Integrated Voltage Measurement X0 = -0.0150000 m, Delta X = 0.0050000 m VT = 0.002952773 Vs 18:13:44 Integrated Voltage Measurement X0 = -0.0150000 m, Delta X = -0.0050000 m VT = -0.002964387 Vs 18:13:44 Integrated Voltage Average and RMS of Measurements X0 = -0.0150000 m, Delta X = 0.0050000 m VT = 0.002959587 +- 0.000001191 Vs 18:13:44 Integrated Field Strength, Average and RMS of Measurements X0 = -0.0150000 m, dx = 0.0050000 m, Nturns = 10.0 VT = 0.002959587 +- 0.000001191 Vs BL = 0.05919174 +- 0.00002382 Tm Remeasure 18:13:44 Integrated Voltage Measurement... 18:13:45 Magnet Current for MAIN, Imag = 154.9751 A Vtrans = 1.549804 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:13:53 Integrated Voltage Measurement X0 = -0.0150000 m, Delta X = 0.0050000 m VT = 0.002952783 Vs 18:14:00 Integrated Voltage Measurement X0 = -0.0150000 m, Delta X = -0.0050000 m VT = -0.002963691 Vs 18:14:01 Magnet Current for MAIN, Imag = 154.9752 A Vtrans = 1.549805 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:14:09 Integrated Voltage Measurement X0 = -0.0150000 m, Delta X = 0.0050000 m VT = 0.002953272 Vs 18:14:16 Integrated Voltage Measurement X0 = -0.0150000 m, Delta X = -0.0050000 m VT = -0.002971085 Vs 18:14:17 Magnet Current for MAIN, Imag = 154.9751 A Vtrans = 1.549804 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:14:24 Integrated Voltage Measurement X0 = -0.0150000 m, Delta X = 0.0050000 m VT = 0.002954551 Vs 18:14:32 Integrated Voltage Measurement X0 = -0.0150000 m, Delta X = -0.0050000 m VT = -0.002966295 Vs 18:14:33 Magnet Current for MAIN, Imag = 154.9752 A Vtrans = 1.549805 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:14:40 Integrated Voltage Measurement X0 = -0.0150000 m, Delta X = 0.0050000 m VT = 0.002953098 Vs 18:14:48 Integrated Voltage Measurement X0 = -0.0150000 m, Delta X = -0.0050000 m VT = -0.002967389 Vs 18:14:48 Integrated Voltage Average and RMS of Measurements X0 = -0.0150000 m, Delta X = 0.0050000 m VT = 0.002960271 +- 0.000001396 Vs 18:14:48 Integrated Field Strength, Average and RMS of Measurements X0 = -0.0150000 m, dx = 0.0050000 m, Nturns = 10.0 VT = 0.002960271 +- 0.000001396 Vs BL = 0.05920541 +- 0.00002793 Tm Remeasure 18:14:48 Integrated Voltage Measurement... 18:14:49 Magnet Current for MAIN, Imag = 154.9750 A Vtrans = 1.549803 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:14:56 Integrated Voltage Measurement X0 = -0.0150000 m, Delta X = 0.0050000 m VT = 0.002948802 Vs 18:15:04 Integrated Voltage Measurement X0 = -0.0150000 m, Delta X = -0.0050000 m VT = -0.002966493 Vs 18:15:04 Magnet Current for MAIN, Imag = 154.9747 A Vtrans = 1.549800 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:15:12 Integrated Voltage Measurement X0 = -0.0150000 m, Delta X = 0.0050000 m VT = 0.002954807 Vs 18:15:20 Integrated Voltage Measurement X0 = -0.0150000 m, Delta X = -0.0050000 m VT = -0.002970512 Vs 18:15:21 Magnet Current for MAIN, Imag = 154.9750 A Vtrans = 1.549803 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:15:28 Integrated Voltage Measurement X0 = -0.0150000 m, Delta X = 0.0050000 m VT = 0.002948573 Vs 18:15:36 Integrated Voltage Measurement X0 = -0.0150000 m, Delta X = -0.0050000 m VT = -0.002965482 Vs 18:15:36 Magnet Current for MAIN, Imag = 154.9748 A Vtrans = 1.549802 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:15:44 Integrated Voltage Measurement X0 = -0.0150000 m, Delta X = 0.0050000 m VT = 0.002952938 Vs 18:15:52 Integrated Voltage Measurement X0 = -0.0150000 m, Delta X = -0.0050000 m VT = -0.002966034 Vs 18:15:52 Integrated Voltage Average and RMS of Measurements X0 = -0.0150000 m, Delta X = 0.0050000 m VT = 0.002959205 +- 0.000002190 Vs 18:15:52 Integrated Field Strength, Average and RMS of Measurements X0 = -0.0150000 m, dx = 0.0050000 m, Nturns = 10.0 VT = 0.002959205 +- 0.000002190 Vs BL = 0.05918411 +- 0.00004380 Tm 18:15:52 Integrated Voltage Measurement... 18:15:52 Magnet Current for MAIN, Imag = 154.9755 A Vtrans = 1.549808 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:16:01 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = 0.0050000 m VT = 0.002957543 Vs 18:16:09 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = -0.0050000 m VT = -0.002975174 Vs 18:16:09 Magnet Current for MAIN, Imag = 154.9750 A Vtrans = 1.549803 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:16:17 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = 0.0050000 m VT = 0.002958068 Vs 18:16:25 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = -0.0050000 m VT = -0.002973856 Vs 18:16:25 Magnet Current for MAIN, Imag = 154.9750 A Vtrans = 1.549803 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:16:33 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = 0.0050000 m VT = 0.002959381 Vs 18:16:41 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = -0.0050000 m VT = -0.002973382 Vs 18:16:41 Magnet Current for MAIN, Imag = 154.9751 A Vtrans = 1.549804 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:16:49 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = 0.0050000 m VT = 0.002955674 Vs 18:16:57 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = -0.0050000 m VT = -0.002975550 Vs 18:16:57 Integrated Voltage Average and RMS of Measurements X0 = -0.0100000 m, Delta X = 0.0050000 m VT = 0.002966079 +- 0.000000317 Vs 18:16:57 Integrated Field Strength, Average and RMS of Measurements X0 = -0.0100000 m, dx = 0.0050000 m, Nturns = 10.0 VT = 0.002966079 +- 0.000000317 Vs BL = 0.05932157 +- 0.00000633 Tm 18:16:57 Integrated Voltage Measurement... 18:16:57 Magnet Current for MAIN, Imag = 154.9751 A Vtrans = 1.549804 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:17:06 Integrated Voltage Measurement X0 = -0.0050000 m, Delta X = 0.0050000 m VT = 0.002962169 Vs 18:17:14 Integrated Voltage Measurement X0 = -0.0050000 m, Delta X = -0.0050000 m VT = -0.002975999 Vs 18:17:14 Magnet Current for MAIN, Imag = 154.9753 A Vtrans = 1.549807 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:17:22 Integrated Voltage Measurement X0 = -0.0050000 m, Delta X = 0.0050000 m VT = 0.002959872 Vs 18:17:29 Integrated Voltage Measurement X0 = -0.0050000 m, Delta X = -0.0050000 m VT = -0.002974264 Vs 18:17:30 Magnet Current for MAIN, Imag = 154.9751 A Vtrans = 1.549804 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:17:38 Integrated Voltage Measurement X0 = -0.0050000 m, Delta X = 0.0050000 m VT = 0.002958799 Vs 18:17:46 Integrated Voltage Measurement X0 = -0.0050000 m, Delta X = -0.0050000 m VT = -0.002974601 Vs 18:17:46 Magnet Current for MAIN, Imag = 154.9752 A Vtrans = 1.549805 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:17:54 Integrated Voltage Measurement X0 = -0.0050000 m, Delta X = 0.0050000 m VT = 0.002956682 Vs 18:18:01 Integrated Voltage Measurement X0 = -0.0050000 m, Delta X = -0.0050000 m VT = -0.002975096 Vs 18:18:01 Integrated Voltage Average and RMS of Measurements X0 = -0.0050000 m, Delta X = 0.0050000 m VT = 0.002967185 +- 0.000001176 Vs 18:18:01 Integrated Field Strength, Average and RMS of Measurements X0 = -0.0050000 m, dx = 0.0050000 m, Nturns = 10.0 VT = 0.002967185 +- 0.000001176 Vs BL = 0.05934371 +- 0.00002352 Tm Remeasure 18:18:01 Integrated Voltage Measurement... 18:18:02 Magnet Current for MAIN, Imag = 154.9752 A Vtrans = 1.549805 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:18:10 Integrated Voltage Measurement X0 = -0.0050000 m, Delta X = 0.0050000 m VT = 0.002959814 Vs 18:18:17 Integrated Voltage Measurement X0 = -0.0050000 m, Delta X = -0.0050000 m VT = -0.002972581 Vs 18:18:18 Magnet Current for MAIN, Imag = 154.9752 A Vtrans = 1.549805 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:18:26 Integrated Voltage Measurement X0 = -0.0050000 m, Delta X = 0.0050000 m VT = 0.002961719 Vs 18:18:33 Integrated Voltage Measurement X0 = -0.0050000 m, Delta X = -0.0050000 m VT = -0.002975255 Vs 18:18:34 Magnet Current for MAIN, Imag = 154.9747 A Vtrans = 1.549800 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:18:42 Integrated Voltage Measurement X0 = -0.0050000 m, Delta X = 0.0050000 m VT = 0.002959437 Vs 18:18:49 Integrated Voltage Measurement X0 = -0.0050000 m, Delta X = -0.0050000 m VT = -0.002973546 Vs 18:18:50 Magnet Current for MAIN, Imag = 154.9752 A Vtrans = 1.549805 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:18:57 Integrated Voltage Measurement X0 = -0.0050000 m, Delta X = 0.0050000 m VT = 0.002957598 Vs 18:19:05 Integrated Voltage Measurement X0 = -0.0050000 m, Delta X = -0.0050000 m VT = -0.002973218 Vs 18:19:05 Integrated Voltage Average and RMS of Measurements X0 = -0.0050000 m, Delta X = 0.0050000 m VT = 0.002966646 +- 0.000001134 Vs 18:19:05 Integrated Field Strength, Average and RMS of Measurements X0 = -0.0050000 m, dx = 0.0050000 m, Nturns = 10.0 VT = 0.002966646 +- 0.000001134 Vs BL = 0.05933292 +- 0.00002269 Tm Remeasure 18:19:05 Integrated Voltage Measurement... 18:19:06 Magnet Current for MAIN, Imag = 154.9751 A Vtrans = 1.549804 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:19:13 Integrated Voltage Measurement X0 = -0.0050000 m, Delta X = 0.0050000 m VT = 0.002959388 Vs 18:19:21 Integrated Voltage Measurement X0 = -0.0050000 m, Delta X = -0.0050000 m VT = -0.002975575 Vs 18:19:21 Magnet Current for MAIN, Imag = 154.9752 A Vtrans = 1.549805 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:19:29 Integrated Voltage Measurement X0 = -0.0050000 m, Delta X = 0.0050000 m VT = 0.002959584 Vs 18:19:36 Integrated Voltage Measurement X0 = -0.0050000 m, Delta X = -0.0050000 m VT = -0.002976858 Vs 18:19:37 Magnet Current for MAIN, Imag = 154.9752 A Vtrans = 1.549805 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:19:45 Integrated Voltage Measurement X0 = -0.0050000 m, Delta X = 0.0050000 m VT = 0.002959352 Vs 18:19:52 Integrated Voltage Measurement X0 = -0.0050000 m, Delta X = -0.0050000 m VT = -0.002975487 Vs 18:19:53 Magnet Current for MAIN, Imag = 154.9751 A Vtrans = 1.549804 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:20:01 Integrated Voltage Measurement X0 = -0.0050000 m, Delta X = 0.0050000 m VT = 0.002960836 Vs 18:20:08 Integrated Voltage Measurement X0 = -0.0050000 m, Delta X = -0.0050000 m VT = -0.002974629 Vs 18:20:08 Integrated Voltage Average and RMS of Measurements X0 = -0.0050000 m, Delta X = 0.0050000 m VT = 0.002967714 +- 0.000000315 Vs 18:20:08 Integrated Field Strength, Average and RMS of Measurements X0 = -0.0050000 m, dx = 0.0050000 m, Nturns = 10.0 VT = 0.002967714 +- 0.000000315 Vs BL = 0.05935427 +- 0.00000631 Tm 18:20:08 Integrated Voltage Measurement... 18:20:09 Magnet Current for MAIN, Imag = 154.9750 A Vtrans = 1.549803 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:20:17 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0050000 m VT = 0.002958256 Vs 18:20:25 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0050000 m VT = -0.002976742 Vs 18:20:26 Magnet Current for MAIN, Imag = 154.9753 A Vtrans = 1.549807 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:20:33 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0050000 m VT = 0.002961023 Vs 18:20:41 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0050000 m VT = -0.002974150 Vs 18:20:42 Magnet Current for MAIN, Imag = 154.9751 A Vtrans = 1.549804 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:20:49 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0050000 m VT = 0.002960460 Vs 18:20:57 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0050000 m VT = -0.002973184 Vs 18:20:58 Magnet Current for MAIN, Imag = 154.9752 A Vtrans = 1.549805 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:21:05 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0050000 m VT = 0.002960898 Vs 18:21:13 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0050000 m VT = -0.002973361 Vs 18:21:13 Integrated Voltage Average and RMS of Measurements X0 = 0.0000000 m, Delta X = 0.0050000 m VT = 0.002967259 +- 0.000000305 Vs 18:21:13 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0000000 m, dx = 0.0050000 m, Nturns = 10.0 VT = 0.002967259 +- 0.000000305 Vs BL = 0.05934518 +- 0.00000611 Tm 18:21:13 Integrated Voltage Measurement... 18:21:14 Magnet Current for MAIN, Imag = 154.9752 A Vtrans = 1.549805 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:21:22 Integrated Voltage Measurement X0 = 0.0050000 m, Delta X = 0.0050000 m VT = 0.002957895 Vs 18:21:30 Integrated Voltage Measurement X0 = 0.0050000 m, Delta X = -0.0050000 m VT = -0.002974823 Vs 18:21:31 Magnet Current for MAIN, Imag = 154.9755 A Vtrans = 1.549808 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:21:38 Integrated Voltage Measurement X0 = 0.0050000 m, Delta X = 0.0050000 m VT = 0.002959954 Vs 18:21:46 Integrated Voltage Measurement X0 = 0.0050000 m, Delta X = -0.0050000 m VT = -0.002973158 Vs 18:21:47 Magnet Current for MAIN, Imag = 154.9751 A Vtrans = 1.549804 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:21:54 Integrated Voltage Measurement X0 = 0.0050000 m, Delta X = 0.0050000 m VT = 0.002956991 Vs 18:22:02 Integrated Voltage Measurement X0 = 0.0050000 m, Delta X = -0.0050000 m VT = -0.002975189 Vs 18:22:02 Magnet Current for MAIN, Imag = 154.9752 A Vtrans = 1.549805 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:22:10 Integrated Voltage Measurement X0 = 0.0050000 m, Delta X = 0.0050000 m VT = 0.002951698 Vs 18:22:18 Integrated Voltage Measurement X0 = 0.0050000 m, Delta X = -0.0050000 m VT = -0.002973163 Vs 18:22:18 Integrated Voltage Average and RMS of Measurements X0 = 0.0050000 m, Delta X = 0.0050000 m VT = 0.002965359 +- 0.000001699 Vs 18:22:18 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0050000 m, dx = 0.0050000 m, Nturns = 10.0 VT = 0.002965359 +- 0.000001699 Vs BL = 0.05930717 +- 0.00003398 Tm Remeasure 18:22:18 Integrated Voltage Measurement... 18:22:18 Magnet Current for MAIN, Imag = 154.9756 A Vtrans = 1.549809 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:22:26 Integrated Voltage Measurement X0 = 0.0050000 m, Delta X = 0.0050000 m VT = 0.002961228 Vs 18:22:33 Integrated Voltage Measurement X0 = 0.0050000 m, Delta X = -0.0050000 m VT = -0.002974580 Vs 18:22:34 Magnet Current for MAIN, Imag = 154.9753 A Vtrans = 1.549807 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:22:42 Integrated Voltage Measurement X0 = 0.0050000 m, Delta X = 0.0050000 m VT = 0.002956821 Vs 18:22:49 Integrated Voltage Measurement X0 = 0.0050000 m, Delta X = -0.0050000 m VT = -0.002974022 Vs 18:22:50 Magnet Current for MAIN, Imag = 154.9752 A Vtrans = 1.549805 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:22:58 Integrated Voltage Measurement X0 = 0.0050000 m, Delta X = 0.0050000 m VT = 0.002959644 Vs 18:23:05 Integrated Voltage Measurement X0 = 0.0050000 m, Delta X = -0.0050000 m VT = -0.002973346 Vs 18:23:06 Magnet Current for MAIN, Imag = 154.9751 A Vtrans = 1.549804 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:23:14 Integrated Voltage Measurement X0 = 0.0050000 m, Delta X = 0.0050000 m VT = 0.002967627 Vs 18:23:21 Integrated Voltage Measurement X0 = 0.0050000 m, Delta X = -0.0050000 m VT = -0.002971955 Vs 18:23:21 Integrated Voltage Average and RMS of Measurements X0 = 0.0050000 m, Delta X = 0.0050000 m VT = 0.002967403 +- 0.000001636 Vs 18:23:21 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0050000 m, dx = 0.0050000 m, Nturns = 10.0 VT = 0.002967403 +- 0.000001636 Vs BL = 0.05934806 +- 0.00003272 Tm Remeasure 18:23:21 Integrated Voltage Measurement... 18:23:22 Magnet Current for MAIN, Imag = 154.9750 A Vtrans = 1.549803 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:23:30 Integrated Voltage Measurement X0 = 0.0050000 m, Delta X = 0.0050000 m VT = 0.002964621 Vs 18:23:38 Integrated Voltage Measurement X0 = 0.0050000 m, Delta X = -0.0050000 m VT = -0.002972198 Vs 18:23:38 Magnet Current for MAIN, Imag = 154.9750 A Vtrans = 1.549803 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:23:46 Integrated Voltage Measurement X0 = 0.0050000 m, Delta X = 0.0050000 m VT = 0.002959722 Vs 18:23:54 Integrated Voltage Measurement X0 = 0.0050000 m, Delta X = -0.0050000 m VT = -0.002971411 Vs 18:23:54 Magnet Current for MAIN, Imag = 154.9757 A Vtrans = 1.549810 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:24:02 Integrated Voltage Measurement X0 = 0.0050000 m, Delta X = 0.0050000 m VT = 0.002944425 Vs 18:24:09 Integrated Voltage Measurement X0 = 0.0050000 m, Delta X = -0.0050000 m VT = -0.002975332 Vs 18:24:10 Magnet Current for MAIN, Imag = 154.9750 A Vtrans = 1.549803 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:24:18 Integrated Voltage Measurement X0 = 0.0050000 m, Delta X = 0.0050000 m VT = 0.002955540 Vs 18:24:25 Integrated Voltage Measurement X0 = 0.0050000 m, Delta X = -0.0050000 m VT = -0.002972822 Vs 18:24:25 Integrated Voltage Average and RMS of Measurements X0 = 0.0050000 m, Delta X = 0.0050000 m VT = 0.002964509 +- 0.000003078 Vs 18:24:25 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0050000 m, dx = 0.0050000 m, Nturns = 10.0 VT = 0.002964509 +- 0.000003078 Vs BL = 0.05929018 +- 0.00006155 Tm Remeasure 18:24:25 Integrated Voltage Measurement... 18:24:26 Magnet Current for MAIN, Imag = 154.9747 A Vtrans = 1.549800 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:24:34 Integrated Voltage Measurement X0 = 0.0050000 m, Delta X = 0.0050000 m VT = 0.002958244 Vs 18:24:41 Integrated Voltage Measurement X0 = 0.0050000 m, Delta X = -0.0050000 m VT = -0.002973342 Vs 18:24:42 Magnet Current for MAIN, Imag = 154.9752 A Vtrans = 1.549805 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:24:50 Integrated Voltage Measurement X0 = 0.0050000 m, Delta X = 0.0050000 m VT = 0.002958270 Vs 18:24:57 Integrated Voltage Measurement X0 = 0.0050000 m, Delta X = -0.0050000 m VT = -0.002972699 Vs 18:24:58 Magnet Current for MAIN, Imag = 154.9750 A Vtrans = 1.549803 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:25:06 Integrated Voltage Measurement X0 = 0.0050000 m, Delta X = 0.0050000 m VT = 0.002958977 Vs 18:25:13 Integrated Voltage Measurement X0 = 0.0050000 m, Delta X = -0.0050000 m VT = -0.002972890 Vs 18:25:14 Magnet Current for MAIN, Imag = 154.9750 A Vtrans = 1.549803 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:25:22 Integrated Voltage Measurement X0 = 0.0050000 m, Delta X = 0.0050000 m VT = 0.002946294 Vs 18:25:29 Integrated Voltage Measurement X0 = 0.0050000 m, Delta X = -0.0050000 m VT = -0.002975046 Vs 18:25:29 Integrated Voltage Average and RMS of Measurements X0 = 0.0050000 m, Delta X = 0.0050000 m VT = 0.002964470 +- 0.000002200 Vs 18:25:29 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0050000 m, dx = 0.0050000 m, Nturns = 10.0 VT = 0.002964470 +- 0.000002200 Vs BL = 0.05928941 +- 0.00004400 Tm Remeasure 18:25:29 Integrated Voltage Measurement... 18:25:30 Magnet Current for MAIN, Imag = 154.9756 A Vtrans = 1.549809 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:25:38 Integrated Voltage Measurement X0 = 0.0050000 m, Delta X = 0.0050000 m VT = 0.002955183 Vs 18:25:45 Integrated Voltage Measurement X0 = 0.0050000 m, Delta X = -0.0050000 m VT = -0.002974685 Vs 18:25:46 Magnet Current for MAIN, Imag = 154.9748 A Vtrans = 1.549802 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:25:53 Integrated Voltage Measurement X0 = 0.0050000 m, Delta X = 0.0050000 m VT = 0.002966472 Vs 18:26:01 Integrated Voltage Measurement X0 = 0.0050000 m, Delta X = -0.0050000 m VT = -0.002972476 Vs 18:26:02 Magnet Current for MAIN, Imag = 154.9756 A Vtrans = 1.549809 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:26:09 Integrated Voltage Measurement X0 = 0.0050000 m, Delta X = 0.0050000 m VT = 0.002967255 Vs 18:26:17 Integrated Voltage Measurement X0 = 0.0050000 m, Delta X = -0.0050000 m VT = -0.002973482 Vs 18:26:18 Magnet Current for MAIN, Imag = 154.9753 A Vtrans = 1.549807 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:26:25 Integrated Voltage Measurement X0 = 0.0050000 m, Delta X = 0.0050000 m VT = 0.002957809 Vs 18:26:33 Integrated Voltage Measurement X0 = 0.0050000 m, Delta X = -0.0050000 m VT = -0.002973470 Vs 18:26:33 Integrated Voltage Average and RMS of Measurements X0 = 0.0050000 m, Delta X = 0.0050000 m VT = 0.002967604 +- 0.000002352 Vs 18:26:33 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0050000 m, dx = 0.0050000 m, Nturns = 10.0 VT = 0.002967604 +- 0.000002352 Vs BL = 0.05935208 +- 0.00004704 Tm 18:26:33 Integrated Voltage Measurement... 18:26:33 Magnet Current for MAIN, Imag = 154.9747 A Vtrans = 1.549800 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:26:42 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = 0.0050000 m VT = 0.002955963 Vs 18:26:50 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = -0.0050000 m VT = -0.002957137 Vs 18:26:50 Magnet Current for MAIN, Imag = 154.9752 A Vtrans = 1.549805 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:26:58 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = 0.0050000 m VT = 0.002954786 Vs 18:27:05 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = -0.0050000 m VT = -0.002965556 Vs 18:27:06 Magnet Current for MAIN, Imag = 154.9748 A Vtrans = 1.549802 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:27:14 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = 0.0050000 m VT = 0.002954349 Vs 18:27:21 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = -0.0050000 m VT = -0.002971138 Vs 18:27:22 Magnet Current for MAIN, Imag = 154.9748 A Vtrans = 1.549802 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:27:30 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = 0.0050000 m VT = 0.002957964 Vs 18:27:37 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = -0.0050000 m VT = -0.002996949 Vs 18:27:37 Integrated Voltage Average and RMS of Measurements X0 = 0.0100000 m, Delta X = 0.0050000 m VT = 0.002964230 +- 0.000007947 Vs 18:27:37 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0100000 m, dx = 0.0050000 m, Nturns = 10.0 VT = 0.002964230 +- 0.000007947 Vs BL = 0.05928461 +- 0.00015894 Tm Remeasure 18:27:37 Integrated Voltage Measurement... 18:27:38 Magnet Current for MAIN, Imag = 154.9752 A Vtrans = 1.549805 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:27:46 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = 0.0050000 m VT = 0.002955931 Vs 18:27:53 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = -0.0050000 m VT = -0.002971632 Vs 18:27:54 Magnet Current for MAIN, Imag = 154.9753 A Vtrans = 1.549807 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:28:02 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = 0.0050000 m VT = 0.002955361 Vs 18:28:09 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = -0.0050000 m VT = -0.002970619 Vs 18:28:10 Magnet Current for MAIN, Imag = 154.9756 A Vtrans = 1.549809 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:28:17 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = 0.0050000 m VT = 0.002961005 Vs 18:28:25 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = -0.0050000 m VT = -0.002975963 Vs 18:28:26 Magnet Current for MAIN, Imag = 154.9746 A Vtrans = 1.549799 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:28:34 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = 0.0050000 m VT = 0.002954561 Vs 18:28:41 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = -0.0050000 m VT = -0.002976109 Vs 18:28:41 Integrated Voltage Average and RMS of Measurements X0 = 0.0100000 m, Delta X = 0.0050000 m VT = 0.002965148 +- 0.000002103 Vs 18:28:41 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0100000 m, dx = 0.0050000 m, Nturns = 10.0 VT = 0.002965148 +- 0.000002103 Vs BL = 0.05930295 +- 0.00004206 Tm Remeasure 18:28:41 Integrated Voltage Measurement... 18:28:42 Magnet Current for MAIN, Imag = 154.9750 A Vtrans = 1.549803 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:28:49 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = 0.0050000 m VT = 0.002955408 Vs 18:28:57 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = -0.0050000 m VT = -0.002973489 Vs 18:28:58 Magnet Current for MAIN, Imag = 154.9747 A Vtrans = 1.549800 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:29:05 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = 0.0050000 m VT = 0.002944662 Vs 18:29:13 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = -0.0050000 m VT = -0.002973608 Vs 18:29:14 Magnet Current for MAIN, Imag = 154.9750 A Vtrans = 1.549803 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:29:21 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = 0.0050000 m VT = 0.002960555 Vs 18:29:29 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = -0.0050000 m VT = -0.002958567 Vs 18:29:30 Magnet Current for MAIN, Imag = 154.9748 A Vtrans = 1.549802 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:29:37 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = 0.0050000 m VT = 0.002961960 Vs 18:29:45 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = -0.0050000 m VT = -0.002981490 Vs 18:29:45 Integrated Voltage Average and RMS of Measurements X0 = 0.0100000 m, Delta X = 0.0050000 m VT = 0.002963717 +- 0.000005073 Vs 18:29:45 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0100000 m, dx = 0.0050000 m, Nturns = 10.0 VT = 0.002963717 +- 0.000005073 Vs BL = 0.05927435 +- 0.00010145 Tm Remeasure 18:29:45 Integrated Voltage Measurement... 18:29:46 Magnet Current for MAIN, Imag = 154.9747 A Vtrans = 1.549800 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:29:53 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = 0.0050000 m VT = 0.002955922 Vs 18:30:01 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = -0.0050000 m VT = -0.002969444 Vs 18:30:02 Magnet Current for MAIN, Imag = 154.9748 A Vtrans = 1.549802 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:30:10 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = 0.0050000 m VT = 0.002956212 Vs 18:30:17 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = -0.0050000 m VT = -0.002985411 Vs 18:30:18 Magnet Current for MAIN, Imag = 154.9750 A Vtrans = 1.549803 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:30:25 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = 0.0050000 m VT = 0.002952653 Vs 18:30:33 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = -0.0050000 m VT = -0.002974420 Vs 18:30:34 Magnet Current for MAIN, Imag = 154.9748 A Vtrans = 1.549802 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:30:41 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = 0.0050000 m VT = 0.002952585 Vs 18:30:49 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = -0.0050000 m VT = -0.002965346 Vs 18:30:49 Integrated Voltage Average and RMS of Measurements X0 = 0.0100000 m, Delta X = 0.0050000 m VT = 0.002963999 +- 0.000004292 Vs 18:30:49 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0100000 m, dx = 0.0050000 m, Nturns = 10.0 VT = 0.002963999 +- 0.000004292 Vs BL = 0.05927998 +- 0.00008585 Tm Remeasure 18:30:49 Integrated Voltage Measurement... 18:30:50 Magnet Current for MAIN, Imag = 154.9752 A Vtrans = 1.549805 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:30:57 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = 0.0050000 m VT = 0.002954572 Vs 18:31:05 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = -0.0050000 m VT = -0.002968445 Vs 18:31:06 Magnet Current for MAIN, Imag = 154.9751 A Vtrans = 1.549804 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:31:13 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = 0.0050000 m VT = 0.002956108 Vs 18:31:21 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = -0.0050000 m VT = -0.002955234 Vs 18:31:22 Magnet Current for MAIN, Imag = 154.9750 A Vtrans = 1.549803 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:31:29 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = 0.0050000 m VT = 0.002953682 Vs 18:31:37 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = -0.0050000 m VT = -0.002972012 Vs 18:31:38 Magnet Current for MAIN, Imag = 154.9751 A Vtrans = 1.549804 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:31:45 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = 0.0050000 m VT = 0.002960586 Vs 18:31:53 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = -0.0050000 m VT = -0.002965471 Vs 18:31:53 Integrated Voltage Average and RMS of Measurements X0 = 0.0100000 m, Delta X = 0.0050000 m VT = 0.002960764 +- 0.000002998 Vs 18:31:53 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0100000 m, dx = 0.0050000 m, Nturns = 10.0 VT = 0.002960764 +- 0.000002998 Vs BL = 0.05921528 +- 0.00005997 Tm 18:31:53 Integrated Voltage Measurement... 18:31:54 Magnet Current for MAIN, Imag = 154.9753 A Vtrans = 1.549807 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:32:02 Integrated Voltage Measurement X0 = 0.0150000 m, Delta X = 0.0050000 m VT = 0.002954545 Vs 18:32:10 Integrated Voltage Measurement X0 = 0.0150000 m, Delta X = -0.0050000 m VT = -0.002960878 Vs 18:32:11 Magnet Current for MAIN, Imag = 154.9747 A Vtrans = 1.549800 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:32:18 Integrated Voltage Measurement X0 = 0.0150000 m, Delta X = 0.0050000 m VT = 0.002950279 Vs 18:32:26 Integrated Voltage Measurement X0 = 0.0150000 m, Delta X = -0.0050000 m VT = -0.002970852 Vs 18:32:27 Magnet Current for MAIN, Imag = 154.9751 A Vtrans = 1.549804 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:32:34 Integrated Voltage Measurement X0 = 0.0150000 m, Delta X = 0.0050000 m VT = 0.002952411 Vs 18:32:42 Integrated Voltage Measurement X0 = 0.0150000 m, Delta X = -0.0050000 m VT = -0.002970861 Vs 18:32:43 Magnet Current for MAIN, Imag = 154.9753 A Vtrans = 1.549807 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:32:50 Integrated Voltage Measurement X0 = 0.0150000 m, Delta X = 0.0050000 m VT = 0.002957526 Vs 18:32:58 Integrated Voltage Measurement X0 = 0.0150000 m, Delta X = -0.0050000 m VT = -0.002964397 Vs 18:32:58 Integrated Voltage Average and RMS of Measurements X0 = 0.0150000 m, Delta X = 0.0050000 m VT = 0.002960219 +- 0.000001497 Vs 18:32:58 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0150000 m, dx = 0.0050000 m, Nturns = 10.0 VT = 0.002960219 +- 0.000001497 Vs BL = 0.05920437 +- 0.00002995 Tm Remeasure 18:32:58 Integrated Voltage Measurement... 18:32:58 Magnet Current for MAIN, Imag = 154.9752 A Vtrans = 1.549805 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:33:06 Integrated Voltage Measurement X0 = 0.0150000 m, Delta X = 0.0050000 m VT = 0.002955382 Vs 18:33:14 Integrated Voltage Measurement X0 = 0.0150000 m, Delta X = -0.0050000 m VT = -0.002966470 Vs 18:33:14 Magnet Current for MAIN, Imag = 154.9750 A Vtrans = 1.549803 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:33:22 Integrated Voltage Measurement X0 = 0.0150000 m, Delta X = 0.0050000 m VT = 0.002951478 Vs 18:33:30 Integrated Voltage Measurement X0 = 0.0150000 m, Delta X = -0.0050000 m VT = -0.002969284 Vs 18:33:30 Magnet Current for MAIN, Imag = 154.9753 A Vtrans = 1.549807 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:33:38 Integrated Voltage Measurement X0 = 0.0150000 m, Delta X = 0.0050000 m VT = 0.002949043 Vs 18:33:45 Integrated Voltage Measurement X0 = 0.0150000 m, Delta X = -0.0050000 m VT = -0.002972742 Vs 18:33:46 Magnet Current for MAIN, Imag = 154.9747 A Vtrans = 1.549800 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:33:54 Integrated Voltage Measurement X0 = 0.0150000 m, Delta X = 0.0050000 m VT = 0.002957923 Vs 18:34:01 Integrated Voltage Measurement X0 = 0.0150000 m, Delta X = -0.0050000 m VT = -0.002965158 Vs 18:34:01 Integrated Voltage Average and RMS of Measurements X0 = 0.0150000 m, Delta X = 0.0050000 m VT = 0.002960935 +- 0.000000411 Vs 18:34:01 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0150000 m, dx = 0.0050000 m, Nturns = 10.0 VT = 0.002960935 +- 0.000000411 Vs BL = 0.05921870 +- 0.00000822 Tm 18:34:01 Integrated Voltage Measurement... 18:34:02 Magnet Current for MAIN, Imag = 154.9750 A Vtrans = 1.549803 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:34:11 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = 0.0050000 m VT = 0.002941100 Vs 18:34:18 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = -0.0050000 m VT = -0.002955419 Vs 18:34:19 Magnet Current for MAIN, Imag = 154.9746 A Vtrans = 1.549799 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:34:27 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = 0.0050000 m VT = 0.002935888 Vs 18:34:34 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = -0.0050000 m VT = -0.002952198 Vs 18:34:35 Magnet Current for MAIN, Imag = 154.9747 A Vtrans = 1.549800 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:34:42 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = 0.0050000 m VT = 0.002938277 Vs 18:34:50 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = -0.0050000 m VT = -0.002954235 Vs 18:34:51 Magnet Current for MAIN, Imag = 154.9752 A Vtrans = 1.549805 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:34:58 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = 0.0050000 m VT = 0.002935677 Vs 18:35:06 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = -0.0050000 m VT = -0.002954896 Vs 18:35:06 Integrated Voltage Average and RMS of Measurements X0 = 0.0200000 m, Delta X = 0.0050000 m VT = 0.002945961 +- 0.000001542 Vs 18:35:06 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0200000 m, dx = 0.0050000 m, Nturns = 10.0 VT = 0.002945961 +- 0.000001542 Vs BL = 0.05891923 +- 0.00003083 Tm Remeasure 18:35:06 Integrated Voltage Measurement... 18:35:07 Magnet Current for MAIN, Imag = 154.9752 A Vtrans = 1.549805 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:35:14 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = 0.0050000 m VT = 0.002937718 Vs 18:35:22 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = -0.0050000 m VT = -0.002952339 Vs 18:35:22 Magnet Current for MAIN, Imag = 154.9752 A Vtrans = 1.549805 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:35:30 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = 0.0050000 m VT = 0.002938657 Vs 18:35:38 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = -0.0050000 m VT = -0.002952709 Vs 18:35:38 Magnet Current for MAIN, Imag = 154.9751 A Vtrans = 1.549804 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:35:46 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = 0.0050000 m VT = 0.002936965 Vs 18:35:54 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = -0.0050000 m VT = -0.002955848 Vs 18:35:54 Magnet Current for MAIN, Imag = 154.9746 A Vtrans = 1.549799 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:36:02 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = 0.0050000 m VT = 0.002939825 Vs 18:36:09 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = -0.0050000 m VT = -0.002954504 Vs 18:36:09 Integrated Voltage Average and RMS of Measurements X0 = 0.0200000 m, Delta X = 0.0050000 m VT = 0.002946071 +- 0.000000798 Vs 18:36:09 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0200000 m, dx = 0.0050000 m, Nturns = 10.0 VT = 0.002946071 +- 0.000000798 Vs BL = 0.05892141 +- 0.00001596 Tm 18:36:09 Integrated Voltage Measurement... 18:36:10 Magnet Current for MAIN, Imag = 154.9745 A Vtrans = 1.549798 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:36:19 Integrated Voltage Measurement X0 = 0.0250000 m, Delta X = 0.0050000 m VT = 0.002893367 Vs 18:36:27 Integrated Voltage Measurement X0 = 0.0250000 m, Delta X = -0.0050000 m VT = -0.002918167 Vs 18:36:27 Magnet Current for MAIN, Imag = 154.9753 A Vtrans = 1.549807 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:36:35 Integrated Voltage Measurement X0 = 0.0250000 m, Delta X = 0.0050000 m VT = 0.002898093 Vs 18:36:42 Integrated Voltage Measurement X0 = 0.0250000 m, Delta X = -0.0050000 m VT = -0.002913085 Vs 18:36:43 Magnet Current for MAIN, Imag = 154.9750 A Vtrans = 1.549803 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:36:51 Integrated Voltage Measurement X0 = 0.0250000 m, Delta X = 0.0050000 m VT = 0.002898251 Vs 18:36:58 Integrated Voltage Measurement X0 = 0.0250000 m, Delta X = -0.0050000 m VT = -0.002916747 Vs 18:36:59 Magnet Current for MAIN, Imag = 154.9748 A Vtrans = 1.549802 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:37:07 Integrated Voltage Measurement X0 = 0.0250000 m, Delta X = 0.0050000 m VT = 0.002899339 Vs 18:37:14 Integrated Voltage Measurement X0 = 0.0250000 m, Delta X = -0.0050000 m VT = -0.002915391 Vs 18:37:14 Integrated Voltage Average and RMS of Measurements X0 = 0.0250000 m, Delta X = 0.0050000 m VT = 0.002906555 +- 0.000000881 Vs 18:37:14 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0250000 m, dx = 0.0050000 m, Nturns = 10.0 VT = 0.002906555 +- 0.000000881 Vs BL = 0.05813110 +- 0.00001761 Tm Remeasure 18:37:14 Integrated Voltage Measurement... 18:37:15 Magnet Current for MAIN, Imag = 154.9751 A Vtrans = 1.549804 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:37:23 Integrated Voltage Measurement X0 = 0.0250000 m, Delta X = 0.0050000 m VT = 0.002901147 Vs 18:37:30 Integrated Voltage Measurement X0 = 0.0250000 m, Delta X = -0.0050000 m VT = -0.002916536 Vs 18:37:31 Magnet Current for MAIN, Imag = 154.9748 A Vtrans = 1.549802 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:37:39 Integrated Voltage Measurement X0 = 0.0250000 m, Delta X = 0.0050000 m VT = 0.002903167 Vs 18:37:46 Integrated Voltage Measurement X0 = 0.0250000 m, Delta X = -0.0050000 m VT = -0.002915154 Vs 18:37:47 Magnet Current for MAIN, Imag = 154.9750 A Vtrans = 1.549803 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:37:54 Integrated Voltage Measurement X0 = 0.0250000 m, Delta X = 0.0050000 m VT = 0.002902155 Vs 18:38:02 Integrated Voltage Measurement X0 = 0.0250000 m, Delta X = -0.0050000 m VT = -0.002913885 Vs 18:38:03 Magnet Current for MAIN, Imag = 154.9747 A Vtrans = 1.549800 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:38:10 Integrated Voltage Measurement X0 = 0.0250000 m, Delta X = 0.0050000 m VT = 0.002903419 Vs 18:38:18 Integrated Voltage Measurement X0 = 0.0250000 m, Delta X = -0.0050000 m VT = -0.002915678 Vs 18:38:18 Integrated Voltage Average and RMS of Measurements X0 = 0.0250000 m, Delta X = 0.0050000 m VT = 0.002908893 +- 0.000000563 Vs 18:38:18 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0250000 m, dx = 0.0050000 m, Nturns = 10.0 VT = 0.002908893 +- 0.000000563 Vs BL = 0.05817786 +- 0.00001125 Tm 18:38:18 Integrated Voltage Measurement... 18:38:19 Magnet Current for MAIN, Imag = 154.9752 A Vtrans = 1.549805 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:38:27 Integrated Voltage Measurement X0 = 0.0300000 m, Delta X = 0.0050000 m VT = 0.002820955 Vs 18:38:35 Integrated Voltage Measurement X0 = 0.0300000 m, Delta X = -0.0050000 m VT = -0.002830626 Vs 18:38:36 Magnet Current for MAIN, Imag = 154.9748 A Vtrans = 1.549802 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:38:43 Integrated Voltage Measurement X0 = 0.0300000 m, Delta X = 0.0050000 m VT = 0.002819706 Vs 18:38:51 Integrated Voltage Measurement X0 = 0.0300000 m, Delta X = -0.0050000 m VT = -0.002834435 Vs 18:38:52 Magnet Current for MAIN, Imag = 154.9746 A Vtrans = 1.549799 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:38:59 Integrated Voltage Measurement X0 = 0.0300000 m, Delta X = 0.0050000 m VT = 0.002823957 Vs 18:39:07 Integrated Voltage Measurement X0 = 0.0300000 m, Delta X = -0.0050000 m VT = -0.002835851 Vs 18:39:07 Magnet Current for MAIN, Imag = 154.9750 A Vtrans = 1.549803 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:39:15 Integrated Voltage Measurement X0 = 0.0300000 m, Delta X = 0.0050000 m VT = 0.002828765 Vs 18:39:23 Integrated Voltage Measurement X0 = 0.0300000 m, Delta X = -0.0050000 m VT = -0.002841308 Vs 18:39:23 Integrated Voltage Average and RMS of Measurements X0 = 0.0300000 m, Delta X = 0.0050000 m VT = 0.002829450 +- 0.000003552 Vs 18:39:23 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0300000 m, dx = 0.0050000 m, Nturns = 10.0 VT = 0.002829450 +- 0.000003552 Vs BL = 0.05658901 +- 0.00007104 Tm Remeasure 18:39:23 Integrated Voltage Measurement... 18:39:23 Magnet Current for MAIN, Imag = 154.9752 A Vtrans = 1.549805 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:39:31 Integrated Voltage Measurement X0 = 0.0300000 m, Delta X = 0.0050000 m VT = 0.002824750 Vs 18:39:39 Integrated Voltage Measurement X0 = 0.0300000 m, Delta X = -0.0050000 m VT = -0.002836825 Vs 18:39:39 Magnet Current for MAIN, Imag = 154.9748 A Vtrans = 1.549802 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:39:47 Integrated Voltage Measurement X0 = 0.0300000 m, Delta X = 0.0050000 m VT = 0.002818077 Vs 18:39:55 Integrated Voltage Measurement X0 = 0.0300000 m, Delta X = -0.0050000 m VT = -0.002834797 Vs 18:39:55 Magnet Current for MAIN, Imag = 154.9751 A Vtrans = 1.549804 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:40:03 Integrated Voltage Measurement X0 = 0.0300000 m, Delta X = 0.0050000 m VT = 0.002822534 Vs 18:40:11 Integrated Voltage Measurement X0 = 0.0300000 m, Delta X = -0.0050000 m VT = -0.002839173 Vs 18:40:11 Magnet Current for MAIN, Imag = 154.9748 A Vtrans = 1.549802 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:40:19 Integrated Voltage Measurement X0 = 0.0300000 m, Delta X = 0.0050000 m VT = 0.002825300 Vs 18:40:26 Integrated Voltage Measurement X0 = 0.0300000 m, Delta X = -0.0050000 m VT = -0.002833594 Vs 18:40:26 Integrated Voltage Average and RMS of Measurements X0 = 0.0300000 m, Delta X = 0.0050000 m VT = 0.002829381 +- 0.000001790 Vs 18:40:26 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0300000 m, dx = 0.0050000 m, Nturns = 10.0 VT = 0.002829381 +- 0.000001790 Vs BL = 0.05658762 +- 0.00003580 Tm Remeasure 18:40:26 Integrated Voltage Measurement... 18:40:27 Magnet Current for MAIN, Imag = 154.9747 A Vtrans = 1.549800 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:40:35 Integrated Voltage Measurement X0 = 0.0300000 m, Delta X = 0.0050000 m VT = 0.002824276 Vs 18:40:42 Integrated Voltage Measurement X0 = 0.0300000 m, Delta X = -0.0050000 m VT = -0.002841093 Vs 18:40:43 Magnet Current for MAIN, Imag = 154.9752 A Vtrans = 1.549805 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:40:51 Integrated Voltage Measurement X0 = 0.0300000 m, Delta X = 0.0050000 m VT = 0.002822088 Vs 18:40:58 Integrated Voltage Measurement X0 = 0.0300000 m, Delta X = -0.0050000 m VT = -0.002836595 Vs 18:40:59 Magnet Current for MAIN, Imag = 154.9748 A Vtrans = 1.549802 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:41:07 Integrated Voltage Measurement X0 = 0.0300000 m, Delta X = 0.0050000 m VT = 0.002823384 Vs 18:41:14 Integrated Voltage Measurement X0 = 0.0300000 m, Delta X = -0.0050000 m VT = -0.002831561 Vs 18:41:15 Magnet Current for MAIN, Imag = 154.9753 A Vtrans = 1.549807 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:41:23 Integrated Voltage Measurement X0 = 0.0300000 m, Delta X = 0.0050000 m VT = 0.002815167 Vs 18:41:30 Integrated Voltage Measurement X0 = 0.0300000 m, Delta X = -0.0050000 m VT = -0.002840694 Vs 18:41:30 Integrated Voltage Average and RMS of Measurements X0 = 0.0300000 m, Delta X = 0.0050000 m VT = 0.002829357 +- 0.000002041 Vs 18:41:30 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0300000 m, dx = 0.0050000 m, Nturns = 10.0 VT = 0.002829357 +- 0.000002041 Vs BL = 0.05658715 +- 0.00004082 Tm Remeasure 18:41:30 Integrated Voltage Measurement... 18:41:31 Magnet Current for MAIN, Imag = 154.9753 A Vtrans = 1.549807 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:41:39 Integrated Voltage Measurement X0 = 0.0300000 m, Delta X = 0.0050000 m VT = 0.002818160 Vs 18:41:46 Integrated Voltage Measurement X0 = 0.0300000 m, Delta X = -0.0050000 m VT = -0.002833908 Vs 18:41:47 Magnet Current for MAIN, Imag = 154.9747 A Vtrans = 1.549800 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:41:55 Integrated Voltage Measurement X0 = 0.0300000 m, Delta X = 0.0050000 m VT = 0.002821797 Vs 18:42:02 Integrated Voltage Measurement X0 = 0.0300000 m, Delta X = -0.0050000 m VT = -0.002837504 Vs 18:42:03 Magnet Current for MAIN, Imag = 154.9747 A Vtrans = 1.549800 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:42:10 Integrated Voltage Measurement X0 = 0.0300000 m, Delta X = 0.0050000 m VT = 0.002817325 Vs 18:42:18 Integrated Voltage Measurement X0 = 0.0300000 m, Delta X = -0.0050000 m VT = -0.002830167 Vs 18:42:19 Magnet Current for MAIN, Imag = 154.9753 A Vtrans = 1.549807 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:42:26 Integrated Voltage Measurement X0 = 0.0300000 m, Delta X = 0.0050000 m VT = 0.002820632 Vs 18:42:34 Integrated Voltage Measurement X0 = 0.0300000 m, Delta X = -0.0050000 m VT = -0.002834813 Vs 18:42:34 Integrated Voltage Average and RMS of Measurements X0 = 0.0300000 m, Delta X = 0.0050000 m VT = 0.002826788 +- 0.000002173 Vs 18:42:34 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0300000 m, dx = 0.0050000 m, Nturns = 10.0 VT = 0.002826788 +- 0.000002173 Vs BL = 0.05653576 +- 0.00004346 Tm Remeasure 18:42:34 Integrated Voltage Measurement... 18:42:35 Magnet Current for MAIN, Imag = 154.9752 A Vtrans = 1.549805 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:42:42 Integrated Voltage Measurement X0 = 0.0300000 m, Delta X = 0.0050000 m VT = 0.002818862 Vs 18:42:50 Integrated Voltage Measurement X0 = 0.0300000 m, Delta X = -0.0050000 m VT = -0.002834784 Vs 18:42:50 Magnet Current for MAIN, Imag = 154.9746 A Vtrans = 1.549799 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:42:58 Integrated Voltage Measurement X0 = 0.0300000 m, Delta X = 0.0050000 m VT = 0.002819339 Vs 18:43:06 Integrated Voltage Measurement X0 = 0.0300000 m, Delta X = -0.0050000 m VT = -0.002839740 Vs 18:43:06 Magnet Current for MAIN, Imag = 154.9751 A Vtrans = 1.549804 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:43:14 Integrated Voltage Measurement X0 = 0.0300000 m, Delta X = 0.0050000 m VT = 0.002820433 Vs 18:43:21 Integrated Voltage Measurement X0 = 0.0300000 m, Delta X = -0.0050000 m VT = -0.002832083 Vs 18:43:22 Magnet Current for MAIN, Imag = 154.9748 A Vtrans = 1.549802 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:43:30 Integrated Voltage Measurement X0 = 0.0300000 m, Delta X = 0.0050000 m VT = 0.002825420 Vs 18:43:37 Integrated Voltage Measurement X0 = 0.0300000 m, Delta X = -0.0050000 m VT = -0.002839401 Vs 18:43:37 Integrated Voltage Average and RMS of Measurements X0 = 0.0300000 m, Delta X = 0.0050000 m VT = 0.002828758 +- 0.000002447 Vs 18:43:37 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0300000 m, dx = 0.0050000 m, Nturns = 10.0 VT = 0.002828758 +- 0.000002447 Vs BL = 0.05657516 +- 0.00004893 Tm 18:43:37 Integrated Voltage Measurement... 18:43:38 Magnet Current for MAIN, Imag = 154.9751 A Vtrans = 1.549804 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:43:47 Integrated Voltage Measurement X0 = 0.0350000 m, Delta X = 0.0050000 m VT = 0.002659952 Vs 18:43:54 Integrated Voltage Measurement X0 = 0.0350000 m, Delta X = -0.0050000 m VT = -0.002687213 Vs 18:43:55 Magnet Current for MAIN, Imag = 154.9753 A Vtrans = 1.549807 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:44:03 Integrated Voltage Measurement X0 = 0.0350000 m, Delta X = 0.0050000 m VT = 0.002660871 Vs 18:44:10 Integrated Voltage Measurement X0 = 0.0350000 m, Delta X = -0.0050000 m VT = -0.002675352 Vs 18:44:11 Magnet Current for MAIN, Imag = 154.9752 A Vtrans = 1.549805 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:44:19 Integrated Voltage Measurement X0 = 0.0350000 m, Delta X = 0.0050000 m VT = 0.002661525 Vs 18:44:26 Integrated Voltage Measurement X0 = 0.0350000 m, Delta X = -0.0050000 m VT = -0.002680014 Vs 18:44:27 Magnet Current for MAIN, Imag = 154.9748 A Vtrans = 1.549802 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:44:35 Integrated Voltage Measurement X0 = 0.0350000 m, Delta X = 0.0050000 m VT = 0.002666298 Vs 18:44:42 Integrated Voltage Measurement X0 = 0.0350000 m, Delta X = -0.0050000 m VT = -0.002674143 Vs 18:44:42 Integrated Voltage Average and RMS of Measurements X0 = 0.0350000 m, Delta X = 0.0050000 m VT = 0.002670671 +- 0.000001952 Vs 18:44:43 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0350000 m, dx = 0.0050000 m, Nturns = 10.0 VT = 0.002670671 +- 0.000001952 Vs BL = 0.05341342 +- 0.00003904 Tm Remeasure 18:44:43 Integrated Voltage Measurement... 18:44:43 Magnet Current for MAIN, Imag = 154.9751 A Vtrans = 1.549804 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:44:51 Integrated Voltage Measurement X0 = 0.0350000 m, Delta X = 0.0050000 m VT = 0.002661585 Vs 18:44:58 Integrated Voltage Measurement X0 = 0.0350000 m, Delta X = -0.0050000 m VT = -0.002681810 Vs 18:44:59 Magnet Current for MAIN, Imag = 154.9747 A Vtrans = 1.549800 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:45:07 Integrated Voltage Measurement X0 = 0.0350000 m, Delta X = 0.0050000 m VT = 0.002645479 Vs 18:45:14 Integrated Voltage Measurement X0 = 0.0350000 m, Delta X = -0.0050000 m VT = -0.002673429 Vs 18:45:15 Magnet Current for MAIN, Imag = 154.9747 A Vtrans = 1.549800 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:45:23 Integrated Voltage Measurement X0 = 0.0350000 m, Delta X = 0.0050000 m VT = 0.002657287 Vs 18:45:30 Integrated Voltage Measurement X0 = 0.0350000 m, Delta X = -0.0050000 m VT = -0.002679872 Vs 18:45:31 Magnet Current for MAIN, Imag = 154.9750 A Vtrans = 1.549803 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:45:39 Integrated Voltage Measurement X0 = 0.0350000 m, Delta X = 0.0050000 m VT = 0.002688906 Vs 18:45:46 Integrated Voltage Measurement X0 = 0.0350000 m, Delta X = -0.0050000 m VT = -0.002680778 Vs 18:45:46 Integrated Voltage Average and RMS of Measurements X0 = 0.0350000 m, Delta X = 0.0050000 m VT = 0.002671143 +- 0.000009099 Vs 18:45:46 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0350000 m, dx = 0.0050000 m, Nturns = 10.0 VT = 0.002671143 +- 0.000009099 Vs BL = 0.05342286 +- 0.00018199 Tm Remeasure 18:45:46 Integrated Voltage Measurement... 18:45:47 Magnet Current for MAIN, Imag = 154.9750 A Vtrans = 1.549803 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:45:54 Integrated Voltage Measurement X0 = 0.0350000 m, Delta X = 0.0050000 m VT = 0.002657839 Vs 18:46:02 Integrated Voltage Measurement X0 = 0.0350000 m, Delta X = -0.0050000 m VT = -0.002677793 Vs 18:46:03 Magnet Current for MAIN, Imag = 154.9748 A Vtrans = 1.549802 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:46:10 Integrated Voltage Measurement X0 = 0.0350000 m, Delta X = 0.0050000 m VT = 0.002668424 Vs 18:46:18 Integrated Voltage Measurement X0 = 0.0350000 m, Delta X = -0.0050000 m VT = -0.002677354 Vs 18:46:19 Magnet Current for MAIN, Imag = 154.9752 A Vtrans = 1.549805 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:46:26 Integrated Voltage Measurement X0 = 0.0350000 m, Delta X = 0.0050000 m VT = 0.002659284 Vs 18:46:34 Integrated Voltage Measurement X0 = 0.0350000 m, Delta X = -0.0050000 m VT = -0.002673039 Vs 18:46:35 Magnet Current for MAIN, Imag = 154.9753 A Vtrans = 1.549807 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:46:42 Integrated Voltage Measurement X0 = 0.0350000 m, Delta X = 0.0050000 m VT = 0.002659584 Vs 18:46:50 Integrated Voltage Measurement X0 = 0.0350000 m, Delta X = -0.0050000 m VT = -0.002682259 Vs 18:46:50 Integrated Voltage Average and RMS of Measurements X0 = 0.0350000 m, Delta X = 0.0050000 m VT = 0.002669447 +- 0.000002621 Vs 18:46:50 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0350000 m, dx = 0.0050000 m, Nturns = 10.0 VT = 0.002669447 +- 0.000002621 Vs BL = 0.05338894 +- 0.00005242 Tm Remeasure 18:46:50 Integrated Voltage Measurement... 18:46:50 Magnet Current for MAIN, Imag = 154.9747 A Vtrans = 1.549800 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:46:58 Integrated Voltage Measurement X0 = 0.0350000 m, Delta X = 0.0050000 m VT = 0.002666285 Vs 18:47:06 Integrated Voltage Measurement X0 = 0.0350000 m, Delta X = -0.0050000 m VT = -0.002677498 Vs 18:47:06 Magnet Current for MAIN, Imag = 154.9748 A Vtrans = 1.549802 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:47:14 Integrated Voltage Measurement X0 = 0.0350000 m, Delta X = 0.0050000 m VT = 0.002682206 Vs 18:47:22 Integrated Voltage Measurement X0 = 0.0350000 m, Delta X = -0.0050000 m VT = -0.002672534 Vs 18:47:22 Magnet Current for MAIN, Imag = 154.9751 A Vtrans = 1.549804 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:47:30 Integrated Voltage Measurement X0 = 0.0350000 m, Delta X = 0.0050000 m VT = 0.002689731 Vs 18:47:38 Integrated Voltage Measurement X0 = 0.0350000 m, Delta X = -0.0050000 m VT = -0.002670694 Vs 18:47:38 Magnet Current for MAIN, Imag = 154.9752 A Vtrans = 1.549805 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:47:46 Integrated Voltage Measurement X0 = 0.0350000 m, Delta X = 0.0050000 m VT = 0.002654897 Vs 18:47:54 Integrated Voltage Measurement X0 = 0.0350000 m, Delta X = -0.0050000 m VT = -0.002676058 Vs 18:47:54 Integrated Voltage Average and RMS of Measurements X0 = 0.0350000 m, Delta X = 0.0050000 m VT = 0.002673738 +- 0.000005629 Vs 18:47:54 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0350000 m, dx = 0.0050000 m, Nturns = 10.0 VT = 0.002673738 +- 0.000005629 Vs BL = 0.05347476 +- 0.00011258 Tm Remeasure 18:47:54 Integrated Voltage Measurement... 18:47:54 Magnet Current for MAIN, Imag = 154.9748 A Vtrans = 1.549802 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:48:02 Integrated Voltage Measurement X0 = 0.0350000 m, Delta X = 0.0050000 m VT = 0.002670235 Vs 18:48:10 Integrated Voltage Measurement X0 = 0.0350000 m, Delta X = -0.0050000 m VT = -0.002675074 Vs 18:48:10 Magnet Current for MAIN, Imag = 154.9746 A Vtrans = 1.549799 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:48:18 Integrated Voltage Measurement X0 = 0.0350000 m, Delta X = 0.0050000 m VT = 0.002672526 Vs 18:48:26 Integrated Voltage Measurement X0 = 0.0350000 m, Delta X = -0.0050000 m VT = -0.002675372 Vs 18:48:26 Magnet Current for MAIN, Imag = 154.9748 A Vtrans = 1.549802 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:48:34 Integrated Voltage Measurement X0 = 0.0350000 m, Delta X = 0.0050000 m VT = 0.002701680 Vs 18:48:41 Integrated Voltage Measurement X0 = 0.0350000 m, Delta X = -0.0050000 m VT = -0.002682879 Vs 18:48:42 Magnet Current for MAIN, Imag = 154.9746 A Vtrans = 1.549799 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:48:50 Integrated Voltage Measurement X0 = 0.0350000 m, Delta X = 0.0050000 m VT = 0.002658250 Vs 18:48:57 Integrated Voltage Measurement X0 = 0.0350000 m, Delta X = -0.0050000 m VT = -0.002679546 Vs 18:48:57 Integrated Voltage Average and RMS of Measurements X0 = 0.0350000 m, Delta X = 0.0050000 m VT = 0.002676945 +- 0.000009046 Vs 18:48:57 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0350000 m, dx = 0.0050000 m, Nturns = 10.0 VT = 0.002676945 +- 0.000009046 Vs BL = 0.05353890 +- 0.00018091 Tm 18:49:01 Position moved: 0.000000 m 18:49:01 Magnet system MAIN current ramping to 155.0000 A... 18:49:12 Magnet Current for MAIN, Imag = 154.9750 A Vtrans = 1.549803 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:49:13 MAIN (V) Voltage Measurement, V = 6.112452 V 18:49:17 Temperature Measurement, Probe Number = 0, T = 26.242000 C 18:49:17 Temperature Measurement, Probe Number = 1, T = 24.344000 C 18:49:18 Temperature Measurement, Probe Number = 2, T = 28.582000 C 18:49:18 Temperature Measurement, Probe Number = 3, T = 30.328000 C 18:49:19 Temperature Measurement, Probe Number = 4, T = 28.665000 C 18:49:20 Temperature Measurement, Probe Number = 5, T = 24.362000 C 18:49:20 Sensor Voltage Measurement, Sensor Number = 0, V = 2.636603 V 18:49:20 Sensor Measurement, Sensor Number = 0, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 2.636603 (V, A, ohm, ...), Measured Value = 122.745255 (C, atm, Hz, ...) 18:49:21 Sensor Voltage Measurement, Sensor Number = 1, V = 1.158211 V 18:49:21 Sensor Measurement, Sensor Number = 1, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 1.158211 (V, A, ohm, ...), Measured Value = 11.865795 (C, atm, Hz, ...) 18:49:21 Magnet Test Parameter Measurements P0 = MAIN, Value = 154.97 P1 = MAIN (V), Value = 6.11 P2 = Yoke, Value = 26.24 P3 = Bottom Coil, Value = 24.34 P4 = Top Coil, Value = 28.58 P5 = Water In, Value = 30.33 P6 = Water Out, Value = 28.66 P7 = Ambient, Value = 24.36 P8 = Supply Pres(psi), Value = 122.75 P9 = Return Pres(psi), Value = 11.87 18:49:21 Integrated Voltage Measurement... 18:49:21 Magnet Current for MAIN, Imag = 154.9746 A Vtrans = 1.549799 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:49:30 Integrated Voltage Measurement X0 = -0.0350000 m, Delta X = 0.0050000 m VT = 0.002657580 Vs 18:49:38 Integrated Voltage Measurement X0 = -0.0350000 m, Delta X = -0.0050000 m VT = -0.002667441 Vs 18:49:38 Magnet Current for MAIN, Imag = 154.9747 A Vtrans = 1.549800 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:49:46 Integrated Voltage Measurement X0 = -0.0350000 m, Delta X = 0.0050000 m VT = 0.002650635 Vs 18:49:54 Integrated Voltage Measurement X0 = -0.0350000 m, Delta X = -0.0050000 m VT = -0.002664039 Vs 18:49:54 Magnet Current for MAIN, Imag = 154.9746 A Vtrans = 1.549799 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:50:02 Integrated Voltage Measurement X0 = -0.0350000 m, Delta X = 0.0050000 m VT = 0.002649513 Vs 18:50:10 Integrated Voltage Measurement X0 = -0.0350000 m, Delta X = -0.0050000 m VT = -0.002670986 Vs 18:50:11 Magnet Current for MAIN, Imag = 154.9752 A Vtrans = 1.549805 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:50:18 Integrated Voltage Measurement X0 = -0.0350000 m, Delta X = 0.0050000 m VT = 0.002652698 Vs 18:50:26 Integrated Voltage Measurement X0 = -0.0350000 m, Delta X = -0.0050000 m VT = -0.002665558 Vs 18:50:26 Integrated Voltage Average and RMS of Measurements X0 = -0.0350000 m, Delta X = 0.0050000 m VT = 0.002659806 +- 0.000001875 Vs 18:50:26 Integrated Field Strength, Average and RMS of Measurements X0 = -0.0350000 m, dx = 0.0050000 m, Nturns = 10.0 VT = 0.002659806 +- 0.000001875 Vs BL = 0.05319613 +- 0.00003751 Tm Remeasure 18:50:26 Integrated Voltage Measurement... 18:50:27 Magnet Current for MAIN, Imag = 154.9746 A Vtrans = 1.549799 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:50:34 Integrated Voltage Measurement X0 = -0.0350000 m, Delta X = 0.0050000 m VT = 0.002658197 Vs 18:50:42 Integrated Voltage Measurement X0 = -0.0350000 m, Delta X = -0.0050000 m VT = -0.002668973 Vs 18:50:42 Magnet Current for MAIN, Imag = 154.9750 A Vtrans = 1.549803 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:50:50 Integrated Voltage Measurement X0 = -0.0350000 m, Delta X = 0.0050000 m VT = 0.002654334 Vs 18:50:57 Integrated Voltage Measurement X0 = -0.0350000 m, Delta X = -0.0050000 m VT = -0.002665225 Vs 18:50:58 Magnet Current for MAIN, Imag = 154.9751 A Vtrans = 1.549804 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:51:06 Integrated Voltage Measurement X0 = -0.0350000 m, Delta X = 0.0050000 m VT = 0.002653494 Vs 18:51:13 Integrated Voltage Measurement X0 = -0.0350000 m, Delta X = -0.0050000 m VT = -0.002665056 Vs 18:51:14 Magnet Current for MAIN, Imag = 154.9748 A Vtrans = 1.549802 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:51:22 Integrated Voltage Measurement X0 = -0.0350000 m, Delta X = 0.0050000 m VT = 0.002652884 Vs 18:51:29 Integrated Voltage Measurement X0 = -0.0350000 m, Delta X = -0.0050000 m VT = -0.002663444 Vs 18:51:29 Integrated Voltage Average and RMS of Measurements X0 = -0.0350000 m, Delta X = 0.0050000 m VT = 0.002660201 +- 0.000002039 Vs 18:51:29 Integrated Field Strength, Average and RMS of Measurements X0 = -0.0350000 m, dx = 0.0050000 m, Nturns = 10.0 VT = 0.002660201 +- 0.000002039 Vs BL = 0.05320401 +- 0.00004079 Tm Remeasure 18:51:29 Integrated Voltage Measurement... 18:51:30 Magnet Current for MAIN, Imag = 154.9747 A Vtrans = 1.549800 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:51:38 Integrated Voltage Measurement X0 = -0.0350000 m, Delta X = 0.0050000 m VT = 0.002654863 Vs 18:51:45 Integrated Voltage Measurement X0 = -0.0350000 m, Delta X = -0.0050000 m VT = -0.002674298 Vs 18:51:46 Magnet Current for MAIN, Imag = 154.9747 A Vtrans = 1.549800 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:51:54 Integrated Voltage Measurement X0 = -0.0350000 m, Delta X = 0.0050000 m VT = 0.002654039 Vs 18:52:01 Integrated Voltage Measurement X0 = -0.0350000 m, Delta X = -0.0050000 m VT = -0.002666172 Vs 18:52:02 Magnet Current for MAIN, Imag = 154.9748 A Vtrans = 1.549802 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:52:10 Integrated Voltage Measurement X0 = -0.0350000 m, Delta X = 0.0050000 m VT = 0.002649800 Vs 18:52:17 Integrated Voltage Measurement X0 = -0.0350000 m, Delta X = -0.0050000 m VT = -0.002672927 Vs 18:52:18 Magnet Current for MAIN, Imag = 154.9748 A Vtrans = 1.549802 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:52:26 Integrated Voltage Measurement X0 = -0.0350000 m, Delta X = 0.0050000 m VT = 0.002652181 Vs 18:52:33 Integrated Voltage Measurement X0 = -0.0350000 m, Delta X = -0.0050000 m VT = -0.002676811 Vs 18:52:33 Integrated Voltage Average and RMS of Measurements X0 = -0.0350000 m, Delta X = 0.0050000 m VT = 0.002662636 +- 0.000001953 Vs 18:52:33 Integrated Field Strength, Average and RMS of Measurements X0 = -0.0350000 m, dx = 0.0050000 m, Nturns = 10.0 VT = 0.002662636 +- 0.000001953 Vs BL = 0.05325272 +- 0.00003907 Tm Remeasure 18:52:33 Integrated Voltage Measurement... 18:52:34 Magnet Current for MAIN, Imag = 154.9750 A Vtrans = 1.549803 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:52:41 Integrated Voltage Measurement X0 = -0.0350000 m, Delta X = 0.0050000 m VT = 0.002655707 Vs 18:52:49 Integrated Voltage Measurement X0 = -0.0350000 m, Delta X = -0.0050000 m VT = -0.002668014 Vs 18:52:50 Magnet Current for MAIN, Imag = 154.9751 A Vtrans = 1.549804 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:52:57 Integrated Voltage Measurement X0 = -0.0350000 m, Delta X = 0.0050000 m VT = 0.002653076 Vs 18:53:05 Integrated Voltage Measurement X0 = -0.0350000 m, Delta X = -0.0050000 m VT = -0.002678284 Vs 18:53:06 Magnet Current for MAIN, Imag = 154.9750 A Vtrans = 1.549803 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:53:13 Integrated Voltage Measurement X0 = -0.0350000 m, Delta X = 0.0050000 m VT = 0.002651628 Vs 18:53:21 Integrated Voltage Measurement X0 = -0.0350000 m, Delta X = -0.0050000 m VT = -0.002666787 Vs 18:53:22 Magnet Current for MAIN, Imag = 154.9746 A Vtrans = 1.549799 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:53:29 Integrated Voltage Measurement X0 = -0.0350000 m, Delta X = 0.0050000 m VT = 0.002654798 Vs 18:53:37 Integrated Voltage Measurement X0 = -0.0350000 m, Delta X = -0.0050000 m VT = -0.002669518 Vs 18:53:37 Integrated Voltage Average and RMS of Measurements X0 = -0.0350000 m, Delta X = 0.0050000 m VT = 0.002662226 +- 0.000002301 Vs 18:53:37 Integrated Field Strength, Average and RMS of Measurements X0 = -0.0350000 m, dx = 0.0050000 m, Nturns = 10.0 VT = 0.002662226 +- 0.000002301 Vs BL = 0.05324453 +- 0.00004602 Tm Remeasure 18:53:37 Integrated Voltage Measurement... 18:53:37 Magnet Current for MAIN, Imag = 154.9750 A Vtrans = 1.549803 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:53:45 Integrated Voltage Measurement X0 = -0.0350000 m, Delta X = 0.0050000 m VT = 0.002654200 Vs 18:53:52 Integrated Voltage Measurement X0 = -0.0350000 m, Delta X = -0.0050000 m VT = -0.002669400 Vs 18:53:53 Magnet Current for MAIN, Imag = 154.9745 A Vtrans = 1.549798 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:54:01 Integrated Voltage Measurement X0 = -0.0350000 m, Delta X = 0.0050000 m VT = 0.002652552 Vs 18:54:08 Integrated Voltage Measurement X0 = -0.0350000 m, Delta X = -0.0050000 m VT = -0.002660118 Vs 18:54:09 Magnet Current for MAIN, Imag = 154.9752 A Vtrans = 1.549805 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:54:17 Integrated Voltage Measurement X0 = -0.0350000 m, Delta X = 0.0050000 m VT = 0.002650218 Vs 18:54:24 Integrated Voltage Measurement X0 = -0.0350000 m, Delta X = -0.0050000 m VT = -0.002664903 Vs 18:54:25 Magnet Current for MAIN, Imag = 154.9748 A Vtrans = 1.549802 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:54:33 Integrated Voltage Measurement X0 = -0.0350000 m, Delta X = 0.0050000 m VT = 0.002649833 Vs 18:54:41 Integrated Voltage Measurement X0 = -0.0350000 m, Delta X = -0.0050000 m VT = -0.002668910 Vs 18:54:41 Integrated Voltage Average and RMS of Measurements X0 = -0.0350000 m, Delta X = 0.0050000 m VT = 0.002658767 +- 0.000002058 Vs 18:54:41 Integrated Field Strength, Average and RMS of Measurements X0 = -0.0350000 m, dx = 0.0050000 m, Nturns = 10.0 VT = 0.002658767 +- 0.000002058 Vs BL = 0.05317534 +- 0.00004115 Tm 18:54:41 Integrated Voltage Measurement... 18:54:41 Magnet Current for MAIN, Imag = 154.9746 A Vtrans = 1.549799 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:54:50 Integrated Voltage Measurement X0 = -0.0300000 m, Delta X = 0.0050000 m VT = 0.002818755 Vs 18:54:58 Integrated Voltage Measurement X0 = -0.0300000 m, Delta X = -0.0050000 m VT = -0.002836372 Vs 18:54:58 Magnet Current for MAIN, Imag = 154.9752 A Vtrans = 1.549805 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:55:06 Integrated Voltage Measurement X0 = -0.0300000 m, Delta X = 0.0050000 m VT = 0.002812872 Vs 18:55:13 Integrated Voltage Measurement X0 = -0.0300000 m, Delta X = -0.0050000 m VT = -0.002835721 Vs 18:55:14 Magnet Current for MAIN, Imag = 154.9750 A Vtrans = 1.549803 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:55:22 Integrated Voltage Measurement X0 = -0.0300000 m, Delta X = 0.0050000 m VT = 0.002814749 Vs 18:55:29 Integrated Voltage Measurement X0 = -0.0300000 m, Delta X = -0.0050000 m VT = -0.002838217 Vs 18:55:30 Magnet Current for MAIN, Imag = 154.9748 A Vtrans = 1.549802 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:55:38 Integrated Voltage Measurement X0 = -0.0300000 m, Delta X = 0.0050000 m VT = 0.002817229 Vs 18:55:45 Integrated Voltage Measurement X0 = -0.0300000 m, Delta X = -0.0050000 m VT = -0.002834907 Vs 18:55:45 Integrated Voltage Average and RMS of Measurements X0 = -0.0300000 m, Delta X = 0.0050000 m VT = 0.002826103 +- 0.000001177 Vs 18:55:45 Integrated Field Strength, Average and RMS of Measurements X0 = -0.0300000 m, dx = 0.0050000 m, Nturns = 10.0 VT = 0.002826103 +- 0.000001177 Vs BL = 0.05652205 +- 0.00002354 Tm Remeasure 18:55:45 Integrated Voltage Measurement... 18:55:46 Magnet Current for MAIN, Imag = 154.9750 A Vtrans = 1.549803 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:55:54 Integrated Voltage Measurement X0 = -0.0300000 m, Delta X = 0.0050000 m VT = 0.002816974 Vs 18:56:01 Integrated Voltage Measurement X0 = -0.0300000 m, Delta X = -0.0050000 m VT = -0.002833408 Vs 18:56:02 Magnet Current for MAIN, Imag = 154.9748 A Vtrans = 1.549802 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:56:09 Integrated Voltage Measurement X0 = -0.0300000 m, Delta X = 0.0050000 m VT = 0.002813569 Vs 18:56:17 Integrated Voltage Measurement X0 = -0.0300000 m, Delta X = -0.0050000 m VT = -0.002834144 Vs 18:56:18 Magnet Current for MAIN, Imag = 154.9750 A Vtrans = 1.549803 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:56:25 Integrated Voltage Measurement X0 = -0.0300000 m, Delta X = 0.0050000 m VT = 0.002815220 Vs 18:56:33 Integrated Voltage Measurement X0 = -0.0300000 m, Delta X = -0.0050000 m VT = -0.002839534 Vs 18:56:34 Magnet Current for MAIN, Imag = 154.9747 A Vtrans = 1.549800 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:56:41 Integrated Voltage Measurement X0 = -0.0300000 m, Delta X = 0.0050000 m VT = 0.002816511 Vs 18:56:49 Integrated Voltage Measurement X0 = -0.0300000 m, Delta X = -0.0050000 m VT = -0.002838651 Vs 18:56:49 Integrated Voltage Average and RMS of Measurements X0 = -0.0300000 m, Delta X = 0.0050000 m VT = 0.002826001 +- 0.000001553 Vs 18:56:49 Integrated Field Strength, Average and RMS of Measurements X0 = -0.0300000 m, dx = 0.0050000 m, Nturns = 10.0 VT = 0.002826001 +- 0.000001553 Vs BL = 0.05652003 +- 0.00003106 Tm Remeasure 18:56:49 Integrated Voltage Measurement... 18:56:49 Magnet Current for MAIN, Imag = 154.9747 A Vtrans = 1.549800 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:56:57 Integrated Voltage Measurement X0 = -0.0300000 m, Delta X = 0.0050000 m VT = 0.002820162 Vs 18:57:04 Integrated Voltage Measurement X0 = -0.0300000 m, Delta X = -0.0050000 m VT = -0.002835606 Vs 18:57:05 Magnet Current for MAIN, Imag = 154.9748 A Vtrans = 1.549802 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:57:13 Integrated Voltage Measurement X0 = -0.0300000 m, Delta X = 0.0050000 m VT = 0.002819773 Vs 18:57:20 Integrated Voltage Measurement X0 = -0.0300000 m, Delta X = -0.0050000 m VT = -0.002827211 Vs 18:57:21 Magnet Current for MAIN, Imag = 154.9752 A Vtrans = 1.549805 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:57:29 Integrated Voltage Measurement X0 = -0.0300000 m, Delta X = 0.0050000 m VT = 0.002816397 Vs 18:57:36 Integrated Voltage Measurement X0 = -0.0300000 m, Delta X = -0.0050000 m VT = -0.002831373 Vs 18:57:37 Magnet Current for MAIN, Imag = 154.9752 A Vtrans = 1.549805 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:57:44 Integrated Voltage Measurement X0 = -0.0300000 m, Delta X = 0.0050000 m VT = 0.002817156 Vs 18:57:52 Integrated Voltage Measurement X0 = -0.0300000 m, Delta X = -0.0050000 m VT = -0.002832380 Vs 18:57:52 Integrated Voltage Average and RMS of Measurements X0 = -0.0300000 m, Delta X = 0.0050000 m VT = 0.002825007 +- 0.000001724 Vs 18:57:52 Integrated Field Strength, Average and RMS of Measurements X0 = -0.0300000 m, dx = 0.0050000 m, Nturns = 10.0 VT = 0.002825007 +- 0.000001724 Vs BL = 0.05650014 +- 0.00003448 Tm Remeasure 18:57:52 Integrated Voltage Measurement... 18:57:53 Magnet Current for MAIN, Imag = 154.9746 A Vtrans = 1.549799 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:58:00 Integrated Voltage Measurement X0 = -0.0300000 m, Delta X = 0.0050000 m VT = 0.002818815 Vs 18:58:08 Integrated Voltage Measurement X0 = -0.0300000 m, Delta X = -0.0050000 m VT = -0.002828511 Vs 18:58:09 Magnet Current for MAIN, Imag = 154.9750 A Vtrans = 1.549803 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:58:16 Integrated Voltage Measurement X0 = -0.0300000 m, Delta X = 0.0050000 m VT = 0.002817671 Vs 18:58:24 Integrated Voltage Measurement X0 = -0.0300000 m, Delta X = -0.0050000 m VT = -0.002832914 Vs 18:58:25 Magnet Current for MAIN, Imag = 154.9751 A Vtrans = 1.549804 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:58:32 Integrated Voltage Measurement X0 = -0.0300000 m, Delta X = 0.0050000 m VT = 0.002818736 Vs 18:58:40 Integrated Voltage Measurement X0 = -0.0300000 m, Delta X = -0.0050000 m VT = -0.002831720 Vs 18:58:41 Magnet Current for MAIN, Imag = 154.9751 A Vtrans = 1.549804 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:58:48 Integrated Voltage Measurement X0 = -0.0300000 m, Delta X = 0.0050000 m VT = 0.002819223 Vs 18:58:56 Integrated Voltage Measurement X0 = -0.0300000 m, Delta X = -0.0050000 m VT = -0.002832692 Vs 18:58:56 Integrated Voltage Average and RMS of Measurements X0 = -0.0300000 m, Delta X = 0.0050000 m VT = 0.002825035 +- 0.000000842 Vs 18:58:56 Integrated Field Strength, Average and RMS of Measurements X0 = -0.0300000 m, dx = 0.0050000 m, Nturns = 10.0 VT = 0.002825035 +- 0.000000842 Vs BL = 0.05650070 +- 0.00001684 Tm 18:58:56 Integrated Voltage Measurement... 18:58:56 Magnet Current for MAIN, Imag = 154.9747 A Vtrans = 1.549800 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:59:05 Integrated Voltage Measurement X0 = -0.0250000 m, Delta X = 0.0050000 m VT = 0.002902441 Vs 18:59:13 Integrated Voltage Measurement X0 = -0.0250000 m, Delta X = -0.0050000 m VT = -0.002914855 Vs 18:59:14 Magnet Current for MAIN, Imag = 154.9746 A Vtrans = 1.549799 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:59:21 Integrated Voltage Measurement X0 = -0.0250000 m, Delta X = 0.0050000 m VT = 0.002897495 Vs 18:59:29 Integrated Voltage Measurement X0 = -0.0250000 m, Delta X = -0.0050000 m VT = -0.002913579 Vs 18:59:29 Magnet Current for MAIN, Imag = 154.9752 A Vtrans = 1.549805 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:59:37 Integrated Voltage Measurement X0 = -0.0250000 m, Delta X = 0.0050000 m VT = 0.002901688 Vs 18:59:45 Integrated Voltage Measurement X0 = -0.0250000 m, Delta X = -0.0050000 m VT = -0.002916796 Vs 18:59:45 Magnet Current for MAIN, Imag = 154.9748 A Vtrans = 1.549802 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 18:59:53 Integrated Voltage Measurement X0 = -0.0250000 m, Delta X = 0.0050000 m VT = 0.002904969 Vs 19:00:00 Integrated Voltage Measurement X0 = -0.0250000 m, Delta X = -0.0050000 m VT = -0.002915231 Vs 19:00:00 Integrated Voltage Average and RMS of Measurements X0 = -0.0250000 m, Delta X = 0.0050000 m VT = 0.002908382 +- 0.000001721 Vs 19:00:00 Integrated Field Strength, Average and RMS of Measurements X0 = -0.0250000 m, dx = 0.0050000 m, Nturns = 10.0 VT = 0.002908382 +- 0.000001721 Vs BL = 0.05816763 +- 0.00003443 Tm Remeasure 19:00:00 Integrated Voltage Measurement... 19:00:01 Magnet Current for MAIN, Imag = 154.9748 A Vtrans = 1.549802 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:00:09 Integrated Voltage Measurement X0 = -0.0250000 m, Delta X = 0.0050000 m VT = 0.002902670 Vs 19:00:16 Integrated Voltage Measurement X0 = -0.0250000 m, Delta X = -0.0050000 m VT = -0.002915339 Vs 19:00:17 Magnet Current for MAIN, Imag = 154.9747 A Vtrans = 1.549800 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:00:24 Integrated Voltage Measurement X0 = -0.0250000 m, Delta X = 0.0050000 m VT = 0.002907206 Vs 19:00:32 Integrated Voltage Measurement X0 = -0.0250000 m, Delta X = -0.0050000 m VT = -0.002915350 Vs 19:00:33 Magnet Current for MAIN, Imag = 154.9745 A Vtrans = 1.549798 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:00:40 Integrated Voltage Measurement X0 = -0.0250000 m, Delta X = 0.0050000 m VT = 0.002901679 Vs 19:00:48 Integrated Voltage Measurement X0 = -0.0250000 m, Delta X = -0.0050000 m VT = -0.002915905 Vs 19:00:49 Magnet Current for MAIN, Imag = 154.9748 A Vtrans = 1.549802 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:00:56 Integrated Voltage Measurement X0 = -0.0250000 m, Delta X = 0.0050000 m VT = 0.002894900 Vs 19:01:04 Integrated Voltage Measurement X0 = -0.0250000 m, Delta X = -0.0050000 m VT = -0.002918980 Vs 19:01:04 Integrated Voltage Average and RMS of Measurements X0 = -0.0250000 m, Delta X = 0.0050000 m VT = 0.002909004 +- 0.000001539 Vs 19:01:04 Integrated Field Strength, Average and RMS of Measurements X0 = -0.0250000 m, dx = 0.0050000 m, Nturns = 10.0 VT = 0.002909004 +- 0.000001539 Vs BL = 0.05818007 +- 0.00003078 Tm Remeasure 19:01:04 Integrated Voltage Measurement... 19:01:04 Magnet Current for MAIN, Imag = 154.9751 A Vtrans = 1.549804 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:01:12 Integrated Voltage Measurement X0 = -0.0250000 m, Delta X = 0.0050000 m VT = 0.002901521 Vs 19:01:20 Integrated Voltage Measurement X0 = -0.0250000 m, Delta X = -0.0050000 m VT = -0.002912472 Vs 19:01:21 Magnet Current for MAIN, Imag = 154.9745 A Vtrans = 1.549798 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:01:28 Integrated Voltage Measurement X0 = -0.0250000 m, Delta X = 0.0050000 m VT = 0.002898683 Vs 19:01:36 Integrated Voltage Measurement X0 = -0.0250000 m, Delta X = -0.0050000 m VT = -0.002914144 Vs 19:01:36 Magnet Current for MAIN, Imag = 154.9748 A Vtrans = 1.549802 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:01:44 Integrated Voltage Measurement X0 = -0.0250000 m, Delta X = 0.0050000 m VT = 0.002899571 Vs 19:01:52 Integrated Voltage Measurement X0 = -0.0250000 m, Delta X = -0.0050000 m VT = -0.002915177 Vs 19:01:52 Magnet Current for MAIN, Imag = 154.9750 A Vtrans = 1.549803 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:02:00 Integrated Voltage Measurement X0 = -0.0250000 m, Delta X = 0.0050000 m VT = 0.002901599 Vs 19:02:07 Integrated Voltage Measurement X0 = -0.0250000 m, Delta X = -0.0050000 m VT = -0.002913345 Vs 19:02:07 Integrated Voltage Average and RMS of Measurements X0 = -0.0250000 m, Delta X = 0.0050000 m VT = 0.002907064 +- 0.000000415 Vs 19:02:07 Integrated Field Strength, Average and RMS of Measurements X0 = -0.0250000 m, dx = 0.0050000 m, Nturns = 10.0 VT = 0.002907064 +- 0.000000415 Vs BL = 0.05814128 +- 0.00000831 Tm 19:02:07 Integrated Voltage Measurement... 19:02:08 Magnet Current for MAIN, Imag = 154.9750 A Vtrans = 1.549803 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:02:17 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = 0.0050000 m VT = 0.002929542 Vs 19:02:24 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = -0.0050000 m VT = -0.002948677 Vs 19:02:25 Magnet Current for MAIN, Imag = 154.9748 A Vtrans = 1.549802 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:02:33 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = 0.0050000 m VT = 0.002937670 Vs 19:02:40 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = -0.0050000 m VT = -0.002959935 Vs 19:02:41 Magnet Current for MAIN, Imag = 154.9746 A Vtrans = 1.549799 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:02:49 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = 0.0050000 m VT = 0.002939094 Vs 19:02:57 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = -0.0050000 m VT = -0.002953996 Vs 19:02:57 Magnet Current for MAIN, Imag = 154.9745 A Vtrans = 1.549798 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:03:05 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = 0.0050000 m VT = 0.002934241 Vs 19:03:12 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = -0.0050000 m VT = -0.002955887 Vs 19:03:12 Integrated Voltage Average and RMS of Measurements X0 = -0.0200000 m, Delta X = 0.0050000 m VT = 0.002944880 +- 0.000003588 Vs 19:03:12 Integrated Field Strength, Average and RMS of Measurements X0 = -0.0200000 m, dx = 0.0050000 m, Nturns = 10.0 VT = 0.002944880 +- 0.000003588 Vs BL = 0.05889760 +- 0.00007176 Tm Remeasure 19:03:12 Integrated Voltage Measurement... 19:03:13 Magnet Current for MAIN, Imag = 154.9752 A Vtrans = 1.549805 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:03:21 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = 0.0050000 m VT = 0.002941028 Vs 19:03:28 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = -0.0050000 m VT = -0.002946791 Vs 19:03:29 Magnet Current for MAIN, Imag = 154.9748 A Vtrans = 1.549802 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:03:37 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = 0.0050000 m VT = 0.002936560 Vs 19:03:44 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = -0.0050000 m VT = -0.002948565 Vs 19:03:45 Magnet Current for MAIN, Imag = 154.9746 A Vtrans = 1.549799 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:03:53 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = 0.0050000 m VT = 0.002937974 Vs 19:04:00 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = -0.0050000 m VT = -0.002943397 Vs 19:04:01 Magnet Current for MAIN, Imag = 154.9750 A Vtrans = 1.549803 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:04:09 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = 0.0050000 m VT = 0.002937066 Vs 19:04:16 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = -0.0050000 m VT = -0.002953335 Vs 19:04:16 Integrated Voltage Average and RMS of Measurements X0 = -0.0200000 m, Delta X = 0.0050000 m VT = 0.002943090 +- 0.000001672 Vs 19:04:16 Integrated Field Strength, Average and RMS of Measurements X0 = -0.0200000 m, dx = 0.0050000 m, Nturns = 10.0 VT = 0.002943090 +- 0.000001672 Vs BL = 0.05886179 +- 0.00003345 Tm Remeasure 19:04:16 Integrated Voltage Measurement... 19:04:17 Magnet Current for MAIN, Imag = 154.9750 A Vtrans = 1.549803 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:04:24 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = 0.0050000 m VT = 0.002939960 Vs 19:04:32 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = -0.0050000 m VT = -0.002952317 Vs 19:04:33 Magnet Current for MAIN, Imag = 154.9751 A Vtrans = 1.549804 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:04:41 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = 0.0050000 m VT = 0.002942507 Vs 19:04:48 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = -0.0050000 m VT = -0.002946677 Vs 19:04:49 Magnet Current for MAIN, Imag = 154.9746 A Vtrans = 1.549799 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:04:56 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = 0.0050000 m VT = 0.002939700 Vs 19:05:04 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = -0.0050000 m VT = -0.002951267 Vs 19:05:05 Magnet Current for MAIN, Imag = 154.9751 A Vtrans = 1.549804 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:05:13 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = 0.0050000 m VT = 0.002939676 Vs 19:05:20 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = -0.0050000 m VT = -0.002955439 Vs 19:05:20 Integrated Voltage Average and RMS of Measurements X0 = -0.0200000 m, Delta X = 0.0050000 m VT = 0.002945943 +- 0.000001082 Vs 19:05:20 Integrated Field Strength, Average and RMS of Measurements X0 = -0.0200000 m, dx = 0.0050000 m, Nturns = 10.0 VT = 0.002945943 +- 0.000001082 Vs BL = 0.05891886 +- 0.00002163 Tm Remeasure 19:05:20 Integrated Voltage Measurement... 19:05:21 Magnet Current for MAIN, Imag = 154.9752 A Vtrans = 1.549805 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:05:29 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = 0.0050000 m VT = 0.002936813 Vs 19:05:36 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = -0.0050000 m VT = -0.002953668 Vs 19:05:37 Magnet Current for MAIN, Imag = 154.9750 A Vtrans = 1.549803 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:05:45 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = 0.0050000 m VT = 0.002937165 Vs 19:05:52 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = -0.0050000 m VT = -0.002951860 Vs 19:05:53 Magnet Current for MAIN, Imag = 154.9750 A Vtrans = 1.549803 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:06:00 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = 0.0050000 m VT = 0.002936401 Vs 19:06:08 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = -0.0050000 m VT = -0.002956144 Vs 19:06:09 Magnet Current for MAIN, Imag = 154.9751 A Vtrans = 1.549804 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:06:16 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = 0.0050000 m VT = 0.002942739 Vs 19:06:24 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = -0.0050000 m VT = -0.002953976 Vs 19:06:24 Integrated Voltage Average and RMS of Measurements X0 = -0.0200000 m, Delta X = 0.0050000 m VT = 0.002946096 +- 0.000001448 Vs 19:06:24 Integrated Field Strength, Average and RMS of Measurements X0 = -0.0200000 m, dx = 0.0050000 m, Nturns = 10.0 VT = 0.002946096 +- 0.000001448 Vs BL = 0.05892191 +- 0.00002896 Tm Remeasure 19:06:24 Integrated Voltage Measurement... 19:06:25 Magnet Current for MAIN, Imag = 154.9752 A Vtrans = 1.549805 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:06:32 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = 0.0050000 m VT = 0.002937039 Vs 19:06:40 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = -0.0050000 m VT = -0.002961285 Vs 19:06:41 Magnet Current for MAIN, Imag = 154.9747 A Vtrans = 1.549800 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:06:48 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = 0.0050000 m VT = 0.002938678 Vs 19:06:56 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = -0.0050000 m VT = -0.002947121 Vs 19:06:57 Magnet Current for MAIN, Imag = 154.9750 A Vtrans = 1.549803 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:07:04 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = 0.0050000 m VT = 0.002938464 Vs 19:07:12 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = -0.0050000 m VT = -0.002958745 Vs 19:07:12 Magnet Current for MAIN, Imag = 154.9748 A Vtrans = 1.549802 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:07:20 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = 0.0050000 m VT = 0.002936170 Vs 19:07:28 Integrated Voltage Measurement X0 = -0.0200000 m, Delta X = -0.0050000 m VT = -0.002957872 Vs 19:07:28 Integrated Voltage Average and RMS of Measurements X0 = -0.0200000 m, Delta X = 0.0050000 m VT = 0.002946922 +- 0.000002452 Vs 19:07:28 Integrated Field Strength, Average and RMS of Measurements X0 = -0.0200000 m, dx = 0.0050000 m, Nturns = 10.0 VT = 0.002946922 +- 0.000002452 Vs BL = 0.05893843 +- 0.00004903 Tm 19:07:28 Integrated Voltage Measurement... 19:07:28 Magnet Current for MAIN, Imag = 154.9746 A Vtrans = 1.549799 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:07:37 Integrated Voltage Measurement X0 = -0.0150000 m, Delta X = 0.0050000 m VT = 0.002952581 Vs 19:07:45 Integrated Voltage Measurement X0 = -0.0150000 m, Delta X = -0.0050000 m VT = -0.002969931 Vs 19:07:45 Magnet Current for MAIN, Imag = 154.9751 A Vtrans = 1.549804 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:07:53 Integrated Voltage Measurement X0 = -0.0150000 m, Delta X = 0.0050000 m VT = 0.002951377 Vs 19:08:01 Integrated Voltage Measurement X0 = -0.0150000 m, Delta X = -0.0050000 m VT = -0.002964664 Vs 19:08:02 Magnet Current for MAIN, Imag = 154.9750 A Vtrans = 1.549803 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:08:09 Integrated Voltage Measurement X0 = -0.0150000 m, Delta X = 0.0050000 m VT = 0.002952252 Vs 19:08:17 Integrated Voltage Measurement X0 = -0.0150000 m, Delta X = -0.0050000 m VT = -0.002969527 Vs 19:08:17 Magnet Current for MAIN, Imag = 154.9752 A Vtrans = 1.549805 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:08:25 Integrated Voltage Measurement X0 = -0.0150000 m, Delta X = 0.0050000 m VT = 0.002953279 Vs 19:08:33 Integrated Voltage Measurement X0 = -0.0150000 m, Delta X = -0.0050000 m VT = -0.002966721 Vs 19:08:33 Integrated Voltage Average and RMS of Measurements X0 = -0.0150000 m, Delta X = 0.0050000 m VT = 0.002960041 +- 0.000001253 Vs 19:08:33 Integrated Field Strength, Average and RMS of Measurements X0 = -0.0150000 m, dx = 0.0050000 m, Nturns = 10.0 VT = 0.002960041 +- 0.000001253 Vs BL = 0.05920083 +- 0.00002506 Tm Remeasure 19:08:33 Integrated Voltage Measurement... 19:08:33 Magnet Current for MAIN, Imag = 154.9746 A Vtrans = 1.549799 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:08:41 Integrated Voltage Measurement X0 = -0.0150000 m, Delta X = 0.0050000 m VT = 0.002952303 Vs 19:08:49 Integrated Voltage Measurement X0 = -0.0150000 m, Delta X = -0.0050000 m VT = -0.002967755 Vs 19:08:49 Magnet Current for MAIN, Imag = 154.9750 A Vtrans = 1.549803 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:08:57 Integrated Voltage Measurement X0 = -0.0150000 m, Delta X = 0.0050000 m VT = 0.002949949 Vs 19:09:04 Integrated Voltage Measurement X0 = -0.0150000 m, Delta X = -0.0050000 m VT = -0.002968018 Vs 19:09:05 Magnet Current for MAIN, Imag = 154.9750 A Vtrans = 1.549803 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:09:13 Integrated Voltage Measurement X0 = -0.0150000 m, Delta X = 0.0050000 m VT = 0.002952082 Vs 19:09:20 Integrated Voltage Measurement X0 = -0.0150000 m, Delta X = -0.0050000 m VT = -0.002967276 Vs 19:09:21 Magnet Current for MAIN, Imag = 154.9751 A Vtrans = 1.549804 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:09:29 Integrated Voltage Measurement X0 = -0.0150000 m, Delta X = 0.0050000 m VT = 0.002952220 Vs 19:09:36 Integrated Voltage Measurement X0 = -0.0150000 m, Delta X = -0.0050000 m VT = -0.002967485 Vs 19:09:36 Integrated Voltage Average and RMS of Measurements X0 = -0.0150000 m, Delta X = 0.0050000 m VT = 0.002959636 +- 0.000000397 Vs 19:09:36 Integrated Field Strength, Average and RMS of Measurements X0 = -0.0150000 m, dx = 0.0050000 m, Nturns = 10.0 VT = 0.002959636 +- 0.000000397 Vs BL = 0.05919272 +- 0.00000793 Tm 19:09:36 Integrated Voltage Measurement... 19:09:37 Magnet Current for MAIN, Imag = 154.9750 A Vtrans = 1.549803 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:09:46 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = 0.0050000 m VT = 0.002958231 Vs 19:09:53 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = -0.0050000 m VT = -0.002976791 Vs 19:09:54 Magnet Current for MAIN, Imag = 154.9751 A Vtrans = 1.549804 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:10:02 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = 0.0050000 m VT = 0.002959591 Vs 19:10:09 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = -0.0050000 m VT = -0.002973911 Vs 19:10:10 Magnet Current for MAIN, Imag = 154.9745 A Vtrans = 1.549798 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:10:18 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = 0.0050000 m VT = 0.002958329 Vs 19:10:25 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = -0.0050000 m VT = -0.002973174 Vs 19:10:26 Magnet Current for MAIN, Imag = 154.9748 A Vtrans = 1.549802 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:10:33 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = 0.0050000 m VT = 0.002958641 Vs 19:10:41 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = -0.0050000 m VT = -0.002969189 Vs 19:10:41 Integrated Voltage Average and RMS of Measurements X0 = -0.0100000 m, Delta X = 0.0050000 m VT = 0.002965982 +- 0.000001347 Vs 19:10:41 Integrated Field Strength, Average and RMS of Measurements X0 = -0.0100000 m, dx = 0.0050000 m, Nturns = 10.0 VT = 0.002965982 +- 0.000001347 Vs BL = 0.05931964 +- 0.00002694 Tm Remeasure 19:10:41 Integrated Voltage Measurement... 19:10:42 Magnet Current for MAIN, Imag = 154.9750 A Vtrans = 1.549803 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:10:49 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = 0.0050000 m VT = 0.002956788 Vs 19:10:57 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = -0.0050000 m VT = -0.002972255 Vs 19:10:58 Magnet Current for MAIN, Imag = 154.9745 A Vtrans = 1.549798 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:11:05 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = 0.0050000 m VT = 0.002959338 Vs 19:11:13 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = -0.0050000 m VT = -0.002973978 Vs 19:11:14 Magnet Current for MAIN, Imag = 154.9745 A Vtrans = 1.549798 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:11:21 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = 0.0050000 m VT = 0.002960684 Vs 19:11:29 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = -0.0050000 m VT = -0.002969413 Vs 19:11:29 Magnet Current for MAIN, Imag = 154.9752 A Vtrans = 1.549805 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:11:37 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = 0.0050000 m VT = 0.002960796 Vs 19:11:45 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = -0.0050000 m VT = -0.002973775 Vs 19:11:45 Integrated Voltage Average and RMS of Measurements X0 = -0.0100000 m, Delta X = 0.0050000 m VT = 0.002965878 +- 0.000001131 Vs 19:11:45 Integrated Field Strength, Average and RMS of Measurements X0 = -0.0100000 m, dx = 0.0050000 m, Nturns = 10.0 VT = 0.002965878 +- 0.000001131 Vs BL = 0.05931756 +- 0.00002262 Tm Remeasure 19:11:45 Integrated Voltage Measurement... 19:11:45 Magnet Current for MAIN, Imag = 154.9751 A Vtrans = 1.549804 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:11:53 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = 0.0050000 m VT = 0.002958635 Vs 19:12:01 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = -0.0050000 m VT = -0.002972275 Vs 19:12:01 Magnet Current for MAIN, Imag = 154.9751 A Vtrans = 1.549804 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:12:09 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = 0.0050000 m VT = 0.002957832 Vs 19:12:17 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = -0.0050000 m VT = -0.002974130 Vs 19:12:17 Magnet Current for MAIN, Imag = 154.9746 A Vtrans = 1.549799 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:12:25 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = 0.0050000 m VT = 0.002958933 Vs 19:12:33 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = -0.0050000 m VT = -0.002972037 Vs 19:12:33 Magnet Current for MAIN, Imag = 154.9745 A Vtrans = 1.549798 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:12:41 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = 0.0050000 m VT = 0.002958597 Vs 19:12:49 Integrated Voltage Measurement X0 = -0.0100000 m, Delta X = -0.0050000 m VT = -0.002974282 Vs 19:12:49 Integrated Voltage Average and RMS of Measurements X0 = -0.0100000 m, Delta X = 0.0050000 m VT = 0.002965840 +- 0.000000404 Vs 19:12:49 Integrated Field Strength, Average and RMS of Measurements X0 = -0.0100000 m, dx = 0.0050000 m, Nturns = 10.0 VT = 0.002965840 +- 0.000000404 Vs BL = 0.05931680 +- 0.00000809 Tm 19:12:49 Integrated Voltage Measurement... 19:12:49 Magnet Current for MAIN, Imag = 154.9747 A Vtrans = 1.549800 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:12:58 Integrated Voltage Measurement X0 = -0.0050000 m, Delta X = 0.0050000 m VT = 0.002959683 Vs 19:13:06 Integrated Voltage Measurement X0 = -0.0050000 m, Delta X = -0.0050000 m VT = -0.002974292 Vs 19:13:06 Magnet Current for MAIN, Imag = 154.9747 A Vtrans = 1.549800 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:13:14 Integrated Voltage Measurement X0 = -0.0050000 m, Delta X = 0.0050000 m VT = 0.002957735 Vs 19:13:22 Integrated Voltage Measurement X0 = -0.0050000 m, Delta X = -0.0050000 m VT = -0.002975258 Vs 19:13:22 Magnet Current for MAIN, Imag = 154.9750 A Vtrans = 1.549803 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:13:30 Integrated Voltage Measurement X0 = -0.0050000 m, Delta X = 0.0050000 m VT = 0.002958679 Vs 19:13:37 Integrated Voltage Measurement X0 = -0.0050000 m, Delta X = -0.0050000 m VT = -0.002973483 Vs 19:13:38 Magnet Current for MAIN, Imag = 154.9746 A Vtrans = 1.549799 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:13:46 Integrated Voltage Measurement X0 = -0.0050000 m, Delta X = 0.0050000 m VT = 0.002958009 Vs 19:13:54 Integrated Voltage Measurement X0 = -0.0050000 m, Delta X = -0.0050000 m VT = -0.002975400 Vs 19:13:54 Integrated Voltage Average and RMS of Measurements X0 = -0.0050000 m, Delta X = 0.0050000 m VT = 0.002966567 +- 0.000000330 Vs 19:13:54 Integrated Field Strength, Average and RMS of Measurements X0 = -0.0050000 m, dx = 0.0050000 m, Nturns = 10.0 VT = 0.002966567 +- 0.000000330 Vs BL = 0.05933135 +- 0.00000661 Tm 19:13:54 Integrated Voltage Measurement... 19:13:54 Magnet Current for MAIN, Imag = 154.9746 A Vtrans = 1.549799 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:14:03 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0050000 m VT = 0.002955237 Vs 19:14:11 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0050000 m VT = -0.002975552 Vs 19:14:11 Magnet Current for MAIN, Imag = 154.9747 A Vtrans = 1.549800 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:14:19 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0050000 m VT = 0.002962090 Vs 19:14:26 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0050000 m VT = -0.002973759 Vs 19:14:27 Magnet Current for MAIN, Imag = 154.9746 A Vtrans = 1.549799 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:14:35 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0050000 m VT = 0.002962034 Vs 19:14:42 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0050000 m VT = -0.002977301 Vs 19:14:43 Magnet Current for MAIN, Imag = 154.9746 A Vtrans = 1.549799 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:14:51 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0050000 m VT = 0.002956585 Vs 19:14:58 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0050000 m VT = -0.002976154 Vs 19:14:58 Integrated Voltage Average and RMS of Measurements X0 = 0.0000000 m, Delta X = 0.0050000 m VT = 0.002967339 +- 0.000001619 Vs 19:14:58 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0000000 m, dx = 0.0050000 m, Nturns = 10.0 VT = 0.002967339 +- 0.000001619 Vs BL = 0.05934678 +- 0.00003238 Tm Remeasure 19:14:58 Integrated Voltage Measurement... 19:14:59 Magnet Current for MAIN, Imag = 154.9746 A Vtrans = 1.549799 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:15:07 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0050000 m VT = 0.002957780 Vs 19:15:14 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0050000 m VT = -0.002976704 Vs 19:15:15 Magnet Current for MAIN, Imag = 154.9743 A Vtrans = 1.549797 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:15:23 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0050000 m VT = 0.002967100 Vs 19:15:30 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0050000 m VT = -0.002976487 Vs 19:15:31 Magnet Current for MAIN, Imag = 154.9747 A Vtrans = 1.549800 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:15:39 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0050000 m VT = 0.002962854 Vs 19:15:46 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0050000 m VT = -0.002975498 Vs 19:15:47 Magnet Current for MAIN, Imag = 154.9746 A Vtrans = 1.549799 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:15:55 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0050000 m VT = 0.002958633 Vs 19:16:02 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0050000 m VT = -0.002975119 Vs 19:16:02 Integrated Voltage Average and RMS of Measurements X0 = 0.0000000 m, Delta X = 0.0050000 m VT = 0.002968772 +- 0.000001951 Vs 19:16:02 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0000000 m, dx = 0.0050000 m, Nturns = 10.0 VT = 0.002968772 +- 0.000001951 Vs BL = 0.05937544 +- 0.00003902 Tm Remeasure 19:16:02 Integrated Voltage Measurement... 19:16:03 Magnet Current for MAIN, Imag = 154.9750 A Vtrans = 1.549803 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:16:11 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0050000 m VT = 0.002956721 Vs 19:16:18 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0050000 m VT = -0.002978334 Vs 19:16:19 Magnet Current for MAIN, Imag = 154.9746 A Vtrans = 1.549799 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:16:27 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0050000 m VT = 0.002957171 Vs 19:16:34 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0050000 m VT = -0.002976035 Vs 19:16:35 Magnet Current for MAIN, Imag = 154.9748 A Vtrans = 1.549802 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:16:43 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0050000 m VT = 0.002959310 Vs 19:16:50 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0050000 m VT = -0.002975498 Vs 19:16:51 Magnet Current for MAIN, Imag = 154.9748 A Vtrans = 1.549802 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:16:59 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0050000 m VT = 0.002959858 Vs 19:17:06 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0050000 m VT = -0.002974261 Vs 19:17:06 Integrated Voltage Average and RMS of Measurements X0 = 0.0000000 m, Delta X = 0.0050000 m VT = 0.002967149 +- 0.000000359 Vs 19:17:06 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0000000 m, dx = 0.0050000 m, Nturns = 10.0 VT = 0.002967149 +- 0.000000359 Vs BL = 0.05934297 +- 0.00000717 Tm 19:17:06 Integrated Voltage Measurement... 19:17:07 Magnet Current for MAIN, Imag = 154.9747 A Vtrans = 1.549800 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:17:16 Integrated Voltage Measurement X0 = 0.0050000 m, Delta X = 0.0050000 m VT = 0.002961385 Vs 19:17:24 Integrated Voltage Measurement X0 = 0.0050000 m, Delta X = -0.0050000 m VT = -0.002976278 Vs 19:17:24 Magnet Current for MAIN, Imag = 154.9748 A Vtrans = 1.549802 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:17:32 Integrated Voltage Measurement X0 = 0.0050000 m, Delta X = 0.0050000 m VT = 0.002954293 Vs 19:17:40 Integrated Voltage Measurement X0 = 0.0050000 m, Delta X = -0.0050000 m VT = -0.002964125 Vs 19:17:40 Magnet Current for MAIN, Imag = 154.9745 A Vtrans = 1.549798 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:17:48 Integrated Voltage Measurement X0 = 0.0050000 m, Delta X = 0.0050000 m VT = 0.002945684 Vs 19:17:55 Integrated Voltage Measurement X0 = 0.0050000 m, Delta X = -0.0050000 m VT = -0.002978107 Vs 19:17:56 Magnet Current for MAIN, Imag = 154.9748 A Vtrans = 1.549802 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:18:04 Integrated Voltage Measurement X0 = 0.0050000 m, Delta X = 0.0050000 m VT = 0.002958046 Vs 19:18:11 Integrated Voltage Measurement X0 = 0.0050000 m, Delta X = -0.0050000 m VT = -0.002973636 Vs 19:18:11 Integrated Voltage Average and RMS of Measurements X0 = 0.0050000 m, Delta X = 0.0050000 m VT = 0.002963944 +- 0.000003678 Vs 19:18:11 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0050000 m, dx = 0.0050000 m, Nturns = 10.0 VT = 0.002963944 +- 0.000003678 Vs BL = 0.05927889 +- 0.00007355 Tm Remeasure 19:18:11 Integrated Voltage Measurement... 19:18:12 Magnet Current for MAIN, Imag = 154.9751 A Vtrans = 1.549804 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:18:20 Integrated Voltage Measurement X0 = 0.0050000 m, Delta X = 0.0050000 m VT = 0.002956310 Vs 19:18:27 Integrated Voltage Measurement X0 = 0.0050000 m, Delta X = -0.0050000 m VT = -0.002975534 Vs 19:18:28 Magnet Current for MAIN, Imag = 154.9746 A Vtrans = 1.549799 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:18:36 Integrated Voltage Measurement X0 = 0.0050000 m, Delta X = 0.0050000 m VT = 0.002962539 Vs 19:18:43 Integrated Voltage Measurement X0 = 0.0050000 m, Delta X = -0.0050000 m VT = -0.002977779 Vs 19:18:44 Magnet Current for MAIN, Imag = 154.9746 A Vtrans = 1.549799 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:18:52 Integrated Voltage Measurement X0 = 0.0050000 m, Delta X = 0.0050000 m VT = 0.002959369 Vs 19:18:59 Integrated Voltage Measurement X0 = 0.0050000 m, Delta X = -0.0050000 m VT = -0.002980271 Vs 19:19:00 Magnet Current for MAIN, Imag = 154.9747 A Vtrans = 1.549800 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:19:08 Integrated Voltage Measurement X0 = 0.0050000 m, Delta X = 0.0050000 m VT = 0.002961104 Vs 19:19:16 Integrated Voltage Measurement X0 = 0.0050000 m, Delta X = -0.0050000 m VT = -0.002973624 Vs 19:19:16 Integrated Voltage Average and RMS of Measurements X0 = 0.0050000 m, Delta X = 0.0050000 m VT = 0.002968316 +- 0.000001753 Vs 19:19:16 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0050000 m, dx = 0.0050000 m, Nturns = 10.0 VT = 0.002968316 +- 0.000001753 Vs BL = 0.05936633 +- 0.00003506 Tm Remeasure 19:19:16 Integrated Voltage Measurement... 19:19:16 Magnet Current for MAIN, Imag = 154.9745 A Vtrans = 1.549798 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:19:24 Integrated Voltage Measurement X0 = 0.0050000 m, Delta X = 0.0050000 m VT = 0.002960550 Vs 19:19:32 Integrated Voltage Measurement X0 = 0.0050000 m, Delta X = -0.0050000 m VT = -0.002974342 Vs 19:19:32 Magnet Current for MAIN, Imag = 154.9748 A Vtrans = 1.549802 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:19:40 Integrated Voltage Measurement X0 = 0.0050000 m, Delta X = 0.0050000 m VT = 0.002955747 Vs 19:19:47 Integrated Voltage Measurement X0 = 0.0050000 m, Delta X = -0.0050000 m VT = -0.002969614 Vs 19:19:48 Magnet Current for MAIN, Imag = 154.9748 A Vtrans = 1.549802 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:19:56 Integrated Voltage Measurement X0 = 0.0050000 m, Delta X = 0.0050000 m VT = 0.002958151 Vs 19:20:03 Integrated Voltage Measurement X0 = 0.0050000 m, Delta X = -0.0050000 m VT = -0.002968139 Vs 19:20:04 Magnet Current for MAIN, Imag = 154.9746 A Vtrans = 1.549799 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:20:12 Integrated Voltage Measurement X0 = 0.0050000 m, Delta X = 0.0050000 m VT = 0.002960425 Vs 19:20:19 Integrated Voltage Measurement X0 = 0.0050000 m, Delta X = -0.0050000 m VT = -0.002971804 Vs 19:20:19 Integrated Voltage Average and RMS of Measurements X0 = 0.0050000 m, Delta X = 0.0050000 m VT = 0.002964846 +- 0.000001997 Vs 19:20:19 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0050000 m, dx = 0.0050000 m, Nturns = 10.0 VT = 0.002964846 +- 0.000001997 Vs BL = 0.05929693 +- 0.00003994 Tm Remeasure 19:20:19 Integrated Voltage Measurement... 19:20:20 Magnet Current for MAIN, Imag = 154.9748 A Vtrans = 1.549802 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:20:28 Integrated Voltage Measurement X0 = 0.0050000 m, Delta X = 0.0050000 m VT = 0.002959140 Vs 19:20:35 Integrated Voltage Measurement X0 = 0.0050000 m, Delta X = -0.0050000 m VT = -0.002975385 Vs 19:20:36 Magnet Current for MAIN, Imag = 154.9742 A Vtrans = 1.549796 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:20:43 Integrated Voltage Measurement X0 = 0.0050000 m, Delta X = 0.0050000 m VT = 0.002955521 Vs 19:20:51 Integrated Voltage Measurement X0 = 0.0050000 m, Delta X = -0.0050000 m VT = -0.002965586 Vs 19:20:52 Magnet Current for MAIN, Imag = 154.9746 A Vtrans = 1.549799 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:20:59 Integrated Voltage Measurement X0 = 0.0050000 m, Delta X = 0.0050000 m VT = 0.002960161 Vs 19:21:07 Integrated Voltage Measurement X0 = 0.0050000 m, Delta X = -0.0050000 m VT = -0.002962946 Vs 19:21:08 Magnet Current for MAIN, Imag = 154.9747 A Vtrans = 1.549800 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:21:15 Integrated Voltage Measurement X0 = 0.0050000 m, Delta X = 0.0050000 m VT = 0.002959736 Vs 19:21:23 Integrated Voltage Measurement X0 = 0.0050000 m, Delta X = -0.0050000 m VT = -0.002970483 Vs 19:21:23 Integrated Voltage Average and RMS of Measurements X0 = 0.0050000 m, Delta X = 0.0050000 m VT = 0.002963620 +- 0.000002700 Vs 19:21:23 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0050000 m, dx = 0.0050000 m, Nturns = 10.0 VT = 0.002963620 +- 0.000002700 Vs BL = 0.05927239 +- 0.00005400 Tm Remeasure 19:21:23 Integrated Voltage Measurement... 19:21:24 Magnet Current for MAIN, Imag = 154.9747 A Vtrans = 1.549800 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:21:31 Integrated Voltage Measurement X0 = 0.0050000 m, Delta X = 0.0050000 m VT = 0.002963197 Vs 19:21:39 Integrated Voltage Measurement X0 = 0.0050000 m, Delta X = -0.0050000 m VT = -0.002976931 Vs 19:21:40 Magnet Current for MAIN, Imag = 154.9746 A Vtrans = 1.549799 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:21:47 Integrated Voltage Measurement X0 = 0.0050000 m, Delta X = 0.0050000 m VT = 0.002976506 Vs 19:21:55 Integrated Voltage Measurement X0 = 0.0050000 m, Delta X = -0.0050000 m VT = -0.002972077 Vs 19:21:56 Magnet Current for MAIN, Imag = 154.9748 A Vtrans = 1.549802 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:22:03 Integrated Voltage Measurement X0 = 0.0050000 m, Delta X = 0.0050000 m VT = 0.002960764 Vs 19:22:11 Integrated Voltage Measurement X0 = 0.0050000 m, Delta X = -0.0050000 m VT = -0.002972632 Vs 19:22:12 Magnet Current for MAIN, Imag = 154.9746 A Vtrans = 1.549799 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:22:19 Integrated Voltage Measurement X0 = 0.0050000 m, Delta X = 0.0050000 m VT = 0.002960073 Vs 19:22:27 Integrated Voltage Measurement X0 = 0.0050000 m, Delta X = -0.0050000 m VT = -0.002979283 Vs 19:22:27 Integrated Voltage Average and RMS of Measurements X0 = 0.0050000 m, Delta X = 0.0050000 m VT = 0.002970183 +- 0.000002706 Vs 19:22:27 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0050000 m, dx = 0.0050000 m, Nturns = 10.0 VT = 0.002970183 +- 0.000002706 Vs BL = 0.05940366 +- 0.00005413 Tm 19:22:27 Integrated Voltage Measurement... 19:22:27 Magnet Current for MAIN, Imag = 154.9748 A Vtrans = 1.549802 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:22:36 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = 0.0050000 m VT = 0.002960324 Vs 19:22:44 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = -0.0050000 m VT = -0.002979287 Vs 19:22:45 Magnet Current for MAIN, Imag = 154.9747 A Vtrans = 1.549800 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:22:52 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = 0.0050000 m VT = 0.002957008 Vs 19:23:00 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = -0.0050000 m VT = -0.002973951 Vs 19:23:01 Magnet Current for MAIN, Imag = 154.9746 A Vtrans = 1.549799 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:23:08 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = 0.0050000 m VT = 0.002954157 Vs 19:23:16 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = -0.0050000 m VT = -0.002950133 Vs 19:23:16 Magnet Current for MAIN, Imag = 154.9746 A Vtrans = 1.549799 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:23:24 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = 0.0050000 m VT = 0.002958122 Vs 19:23:32 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = -0.0050000 m VT = -0.002972128 Vs 19:23:32 Integrated Voltage Average and RMS of Measurements X0 = 0.0100000 m, Delta X = 0.0050000 m VT = 0.002963139 +- 0.000006609 Vs 19:23:32 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0100000 m, dx = 0.0050000 m, Nturns = 10.0 VT = 0.002963139 +- 0.000006609 Vs BL = 0.05926277 +- 0.00013219 Tm Remeasure 19:23:32 Integrated Voltage Measurement... 19:23:32 Magnet Current for MAIN, Imag = 154.9743 A Vtrans = 1.549797 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:23:40 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = 0.0050000 m VT = 0.002959468 Vs 19:23:48 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = -0.0050000 m VT = -0.002973344 Vs 19:23:48 Magnet Current for MAIN, Imag = 154.9747 A Vtrans = 1.549800 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:23:56 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = 0.0050000 m VT = 0.002956155 Vs 19:24:03 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = -0.0050000 m VT = -0.002980161 Vs 19:24:04 Magnet Current for MAIN, Imag = 154.9746 A Vtrans = 1.549799 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:24:12 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = 0.0050000 m VT = 0.002957408 Vs 19:24:19 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = -0.0050000 m VT = -0.002972404 Vs 19:24:20 Magnet Current for MAIN, Imag = 154.9741 A Vtrans = 1.549794 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:24:27 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = 0.0050000 m VT = 0.002957256 Vs 19:24:35 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = -0.0050000 m VT = -0.002962376 Vs 19:24:35 Integrated Voltage Average and RMS of Measurements X0 = 0.0100000 m, Delta X = 0.0050000 m VT = 0.002964821 +- 0.000003111 Vs 19:24:35 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0100000 m, dx = 0.0050000 m, Nturns = 10.0 VT = 0.002964821 +- 0.000003111 Vs BL = 0.05929643 +- 0.00006221 Tm Remeasure 19:24:35 Integrated Voltage Measurement... 19:24:36 Magnet Current for MAIN, Imag = 154.9741 A Vtrans = 1.549794 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:24:43 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = 0.0050000 m VT = 0.002958896 Vs 19:24:51 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = -0.0050000 m VT = -0.002973239 Vs 19:24:52 Magnet Current for MAIN, Imag = 154.9746 A Vtrans = 1.549799 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:24:59 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = 0.0050000 m VT = 0.002954624 Vs 19:25:07 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = -0.0050000 m VT = -0.002975007 Vs 19:25:08 Magnet Current for MAIN, Imag = 154.9747 A Vtrans = 1.549800 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:25:15 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = 0.0050000 m VT = 0.002956640 Vs 19:25:23 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = -0.0050000 m VT = -0.002968724 Vs 19:25:24 Magnet Current for MAIN, Imag = 154.9746 A Vtrans = 1.549799 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:25:31 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = 0.0050000 m VT = 0.002963398 Vs 19:25:39 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = -0.0050000 m VT = -0.002975285 Vs 19:25:39 Integrated Voltage Average and RMS of Measurements X0 = 0.0100000 m, Delta X = 0.0050000 m VT = 0.002965727 +- 0.000002412 Vs 19:25:39 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0100000 m, dx = 0.0050000 m, Nturns = 10.0 VT = 0.002965727 +- 0.000002412 Vs BL = 0.05931453 +- 0.00004825 Tm Remeasure 19:25:39 Integrated Voltage Measurement... 19:25:40 Magnet Current for MAIN, Imag = 154.9743 A Vtrans = 1.549797 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:25:47 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = 0.0050000 m VT = 0.002957309 Vs 19:25:55 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = -0.0050000 m VT = -0.002982017 Vs 19:25:56 Magnet Current for MAIN, Imag = 154.9746 A Vtrans = 1.549799 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:26:03 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = 0.0050000 m VT = 0.002956163 Vs 19:26:11 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = -0.0050000 m VT = -0.002975326 Vs 19:26:11 Magnet Current for MAIN, Imag = 154.9746 A Vtrans = 1.549799 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:26:19 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = 0.0050000 m VT = 0.002956900 Vs 19:26:27 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = -0.0050000 m VT = -0.002973291 Vs 19:26:28 Magnet Current for MAIN, Imag = 154.9748 A Vtrans = 1.549802 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:26:35 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = 0.0050000 m VT = 0.002953942 Vs 19:26:43 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = -0.0050000 m VT = -0.002968293 Vs 19:26:43 Integrated Voltage Average and RMS of Measurements X0 = 0.0100000 m, Delta X = 0.0050000 m VT = 0.002965405 +- 0.000003030 Vs 19:26:43 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0100000 m, dx = 0.0050000 m, Nturns = 10.0 VT = 0.002965405 +- 0.000003030 Vs BL = 0.05930810 +- 0.00006060 Tm Remeasure 19:26:43 Integrated Voltage Measurement... 19:26:44 Magnet Current for MAIN, Imag = 154.9748 A Vtrans = 1.549802 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:26:51 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = 0.0050000 m VT = 0.002959440 Vs 19:26:59 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = -0.0050000 m VT = -0.002971408 Vs 19:26:59 Magnet Current for MAIN, Imag = 154.9747 A Vtrans = 1.549800 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:27:07 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = 0.0050000 m VT = 0.002956953 Vs 19:27:14 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = -0.0050000 m VT = -0.002982461 Vs 19:27:15 Magnet Current for MAIN, Imag = 154.9747 A Vtrans = 1.549800 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:27:23 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = 0.0050000 m VT = 0.002958247 Vs 19:27:30 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = -0.0050000 m VT = -0.002975282 Vs 19:27:31 Magnet Current for MAIN, Imag = 154.9743 A Vtrans = 1.549797 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:27:39 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = 0.0050000 m VT = 0.002957645 Vs 19:27:46 Integrated Voltage Measurement X0 = 0.0100000 m, Delta X = -0.0050000 m VT = -0.002976166 Vs 19:27:46 Integrated Voltage Average and RMS of Measurements X0 = 0.0100000 m, Delta X = 0.0050000 m VT = 0.002967200 +- 0.000001558 Vs 19:27:46 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0100000 m, dx = 0.0050000 m, Nturns = 10.0 VT = 0.002967200 +- 0.000001558 Vs BL = 0.05934400 +- 0.00003117 Tm 19:27:46 Integrated Voltage Measurement... 19:27:47 Magnet Current for MAIN, Imag = 154.9747 A Vtrans = 1.549800 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:27:56 Integrated Voltage Measurement X0 = 0.0150000 m, Delta X = 0.0050000 m VT = 0.002952783 Vs 19:28:03 Integrated Voltage Measurement X0 = 0.0150000 m, Delta X = -0.0050000 m VT = -0.002968756 Vs 19:28:04 Magnet Current for MAIN, Imag = 154.9748 A Vtrans = 1.549802 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:28:12 Integrated Voltage Measurement X0 = 0.0150000 m, Delta X = 0.0050000 m VT = 0.002951913 Vs 19:28:19 Integrated Voltage Measurement X0 = 0.0150000 m, Delta X = -0.0050000 m VT = -0.002968190 Vs 19:28:20 Magnet Current for MAIN, Imag = 154.9748 A Vtrans = 1.549802 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:28:27 Integrated Voltage Measurement X0 = 0.0150000 m, Delta X = 0.0050000 m VT = 0.002953941 Vs 19:28:35 Integrated Voltage Measurement X0 = 0.0150000 m, Delta X = -0.0050000 m VT = -0.002965354 Vs 19:28:36 Magnet Current for MAIN, Imag = 154.9747 A Vtrans = 1.549800 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:28:43 Integrated Voltage Measurement X0 = 0.0150000 m, Delta X = 0.0050000 m VT = 0.002951317 Vs 19:28:51 Integrated Voltage Measurement X0 = 0.0150000 m, Delta X = -0.0050000 m VT = -0.002964455 Vs 19:28:51 Integrated Voltage Average and RMS of Measurements X0 = 0.0150000 m, Delta X = 0.0050000 m VT = 0.002959589 +- 0.000001062 Vs 19:28:51 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0150000 m, dx = 0.0050000 m, Nturns = 10.0 VT = 0.002959589 +- 0.000001062 Vs BL = 0.05919177 +- 0.00002124 Tm Remeasure 19:28:51 Integrated Voltage Measurement... 19:28:52 Magnet Current for MAIN, Imag = 154.9745 A Vtrans = 1.549798 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:28:59 Integrated Voltage Measurement X0 = 0.0150000 m, Delta X = 0.0050000 m VT = 0.002951665 Vs 19:29:07 Integrated Voltage Measurement X0 = 0.0150000 m, Delta X = -0.0050000 m VT = -0.002971653 Vs 19:29:08 Magnet Current for MAIN, Imag = 154.9747 A Vtrans = 1.549800 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:29:15 Integrated Voltage Measurement X0 = 0.0150000 m, Delta X = 0.0050000 m VT = 0.002955753 Vs 19:29:23 Integrated Voltage Measurement X0 = 0.0150000 m, Delta X = -0.0050000 m VT = -0.002966203 Vs 19:29:24 Magnet Current for MAIN, Imag = 154.9743 A Vtrans = 1.549797 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:29:31 Integrated Voltage Measurement X0 = 0.0150000 m, Delta X = 0.0050000 m VT = 0.002951840 Vs 19:29:39 Integrated Voltage Measurement X0 = 0.0150000 m, Delta X = -0.0050000 m VT = -0.002967584 Vs 19:29:40 Magnet Current for MAIN, Imag = 154.9746 A Vtrans = 1.549799 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:29:47 Integrated Voltage Measurement X0 = 0.0150000 m, Delta X = 0.0050000 m VT = 0.002952560 Vs 19:29:55 Integrated Voltage Measurement X0 = 0.0150000 m, Delta X = -0.0050000 m VT = -0.002966768 Vs 19:29:55 Integrated Voltage Average and RMS of Measurements X0 = 0.0150000 m, Delta X = 0.0050000 m VT = 0.002960503 +- 0.000000850 Vs 19:29:55 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0150000 m, dx = 0.0050000 m, Nturns = 10.0 VT = 0.002960503 +- 0.000000850 Vs BL = 0.05921006 +- 0.00001700 Tm 19:29:55 Integrated Voltage Measurement... 19:29:55 Magnet Current for MAIN, Imag = 154.9738 A Vtrans = 1.549792 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:30:04 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = 0.0050000 m VT = 0.002939729 Vs 19:30:12 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = -0.0050000 m VT = -0.002949449 Vs 19:30:12 Magnet Current for MAIN, Imag = 154.9742 A Vtrans = 1.549796 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:30:20 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = 0.0050000 m VT = 0.002938343 Vs 19:30:28 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = -0.0050000 m VT = -0.002952822 Vs 19:30:28 Magnet Current for MAIN, Imag = 154.9743 A Vtrans = 1.549797 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:30:36 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = 0.0050000 m VT = 0.002940886 Vs 19:30:44 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = -0.0050000 m VT = -0.002950343 Vs 19:30:44 Magnet Current for MAIN, Imag = 154.9746 A Vtrans = 1.549799 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:30:52 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = 0.0050000 m VT = 0.002934817 Vs 19:30:59 Integrated Voltage Measurement X0 = 0.0200000 m, Delta X = -0.0050000 m VT = -0.002954560 Vs 19:30:59 Integrated Voltage Average and RMS of Measurements X0 = 0.0200000 m, Delta X = 0.0050000 m VT = 0.002945119 +- 0.000000481 Vs 19:30:59 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0200000 m, dx = 0.0050000 m, Nturns = 10.0 VT = 0.002945119 +- 0.000000481 Vs BL = 0.05890237 +- 0.00000963 Tm 19:31:00 Integrated Voltage Measurement... 19:31:00 Magnet Current for MAIN, Imag = 154.9743 A Vtrans = 1.549797 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:31:09 Integrated Voltage Measurement X0 = 0.0250000 m, Delta X = 0.0050000 m VT = 0.002900145 Vs 19:31:16 Integrated Voltage Measurement X0 = 0.0250000 m, Delta X = -0.0050000 m VT = -0.002911192 Vs 19:31:17 Magnet Current for MAIN, Imag = 154.9742 A Vtrans = 1.549796 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:31:25 Integrated Voltage Measurement X0 = 0.0250000 m, Delta X = 0.0050000 m VT = 0.002892418 Vs 19:31:32 Integrated Voltage Measurement X0 = 0.0250000 m, Delta X = -0.0050000 m VT = -0.002911890 Vs 19:31:33 Magnet Current for MAIN, Imag = 154.9746 A Vtrans = 1.549799 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:31:41 Integrated Voltage Measurement X0 = 0.0250000 m, Delta X = 0.0050000 m VT = 0.002902227 Vs 19:31:48 Integrated Voltage Measurement X0 = 0.0250000 m, Delta X = -0.0050000 m VT = -0.002915448 Vs 19:31:49 Magnet Current for MAIN, Imag = 154.9742 A Vtrans = 1.549796 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:31:56 Integrated Voltage Measurement X0 = 0.0250000 m, Delta X = 0.0050000 m VT = 0.002902562 Vs 19:32:04 Integrated Voltage Measurement X0 = 0.0250000 m, Delta X = -0.0050000 m VT = -0.002911961 Vs 19:32:04 Integrated Voltage Average and RMS of Measurements X0 = 0.0250000 m, Delta X = 0.0050000 m VT = 0.002905980 +- 0.000002477 Vs 19:32:04 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0250000 m, dx = 0.0050000 m, Nturns = 10.0 VT = 0.002905980 +- 0.000002477 Vs BL = 0.05811961 +- 0.00004954 Tm Remeasure 19:32:04 Integrated Voltage Measurement... 19:32:05 Magnet Current for MAIN, Imag = 154.9746 A Vtrans = 1.549799 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:32:12 Integrated Voltage Measurement X0 = 0.0250000 m, Delta X = 0.0050000 m VT = 0.002898093 Vs 19:32:20 Integrated Voltage Measurement X0 = 0.0250000 m, Delta X = -0.0050000 m VT = -0.002913591 Vs 19:32:21 Magnet Current for MAIN, Imag = 154.9746 A Vtrans = 1.549799 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:32:28 Integrated Voltage Measurement X0 = 0.0250000 m, Delta X = 0.0050000 m VT = 0.002900784 Vs 19:32:36 Integrated Voltage Measurement X0 = 0.0250000 m, Delta X = -0.0050000 m VT = -0.002915224 Vs 19:32:36 Magnet Current for MAIN, Imag = 154.9743 A Vtrans = 1.549797 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:32:44 Integrated Voltage Measurement X0 = 0.0250000 m, Delta X = 0.0050000 m VT = 0.002905608 Vs 19:32:52 Integrated Voltage Measurement X0 = 0.0250000 m, Delta X = -0.0050000 m VT = -0.002916139 Vs 19:32:52 Magnet Current for MAIN, Imag = 154.9742 A Vtrans = 1.549796 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:33:00 Integrated Voltage Measurement X0 = 0.0250000 m, Delta X = 0.0050000 m VT = 0.002899043 Vs 19:33:07 Integrated Voltage Measurement X0 = 0.0250000 m, Delta X = -0.0050000 m VT = -0.002915071 Vs 19:33:07 Integrated Voltage Average and RMS of Measurements X0 = 0.0250000 m, Delta X = 0.0050000 m VT = 0.002907944 +- 0.000001857 Vs 19:33:07 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0250000 m, dx = 0.0050000 m, Nturns = 10.0 VT = 0.002907944 +- 0.000001857 Vs BL = 0.05815888 +- 0.00003714 Tm Remeasure 19:33:07 Integrated Voltage Measurement... 19:33:08 Magnet Current for MAIN, Imag = 154.9746 A Vtrans = 1.549799 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:33:16 Integrated Voltage Measurement X0 = 0.0250000 m, Delta X = 0.0050000 m VT = 0.002896851 Vs 19:33:23 Integrated Voltage Measurement X0 = 0.0250000 m, Delta X = -0.0050000 m VT = -0.002919714 Vs 19:33:24 Magnet Current for MAIN, Imag = 154.9747 A Vtrans = 1.549800 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:33:32 Integrated Voltage Measurement X0 = 0.0250000 m, Delta X = 0.0050000 m VT = 0.002901695 Vs 19:33:39 Integrated Voltage Measurement X0 = 0.0250000 m, Delta X = -0.0050000 m VT = -0.002914211 Vs 19:33:40 Magnet Current for MAIN, Imag = 154.9748 A Vtrans = 1.549802 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:33:47 Integrated Voltage Measurement X0 = 0.0250000 m, Delta X = 0.0050000 m VT = 0.002901104 Vs 19:33:55 Integrated Voltage Measurement X0 = 0.0250000 m, Delta X = -0.0050000 m VT = -0.002915292 Vs 19:33:56 Magnet Current for MAIN, Imag = 154.9742 A Vtrans = 1.549796 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:34:03 Integrated Voltage Measurement X0 = 0.0250000 m, Delta X = 0.0050000 m VT = 0.002906224 Vs 19:34:11 Integrated Voltage Measurement X0 = 0.0250000 m, Delta X = -0.0050000 m VT = -0.002914651 Vs 19:34:11 Integrated Voltage Average and RMS of Measurements X0 = 0.0250000 m, Delta X = 0.0050000 m VT = 0.002908718 +- 0.000001000 Vs 19:34:11 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0250000 m, dx = 0.0050000 m, Nturns = 10.0 VT = 0.002908718 +- 0.000001000 Vs BL = 0.05817435 +- 0.00002000 Tm Remeasure 19:34:11 Integrated Voltage Measurement... 19:34:12 Magnet Current for MAIN, Imag = 154.9746 A Vtrans = 1.549799 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:34:19 Integrated Voltage Measurement X0 = 0.0250000 m, Delta X = 0.0050000 m VT = 0.002899516 Vs 19:34:27 Integrated Voltage Measurement X0 = 0.0250000 m, Delta X = -0.0050000 m VT = -0.002918307 Vs 19:34:27 Magnet Current for MAIN, Imag = 154.9745 A Vtrans = 1.549798 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:34:35 Integrated Voltage Measurement X0 = 0.0250000 m, Delta X = 0.0050000 m VT = 0.002896938 Vs 19:34:43 Integrated Voltage Measurement X0 = 0.0250000 m, Delta X = -0.0050000 m VT = -0.002919948 Vs 19:34:43 Magnet Current for MAIN, Imag = 154.9743 A Vtrans = 1.549797 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:34:51 Integrated Voltage Measurement X0 = 0.0250000 m, Delta X = 0.0050000 m VT = 0.002900046 Vs 19:34:58 Integrated Voltage Measurement X0 = 0.0250000 m, Delta X = -0.0050000 m VT = -0.002912426 Vs 19:34:59 Magnet Current for MAIN, Imag = 154.9743 A Vtrans = 1.549797 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:35:07 Integrated Voltage Measurement X0 = 0.0250000 m, Delta X = 0.0050000 m VT = 0.002900262 Vs 19:35:14 Integrated Voltage Measurement X0 = 0.0250000 m, Delta X = -0.0050000 m VT = -0.002912959 Vs 19:35:14 Integrated Voltage Average and RMS of Measurements X0 = 0.0250000 m, Delta X = 0.0050000 m VT = 0.002907550 +- 0.000001147 Vs 19:35:14 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0250000 m, dx = 0.0050000 m, Nturns = 10.0 VT = 0.002907550 +- 0.000001147 Vs BL = 0.05815100 +- 0.00002294 Tm 19:35:14 Integrated Voltage Measurement... 19:35:15 Magnet Current for MAIN, Imag = 154.9743 A Vtrans = 1.549797 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:35:24 Integrated Voltage Measurement X0 = 0.0300000 m, Delta X = 0.0050000 m VT = 0.002830451 Vs 19:35:32 Integrated Voltage Measurement X0 = 0.0300000 m, Delta X = -0.0050000 m VT = -0.002840407 Vs 19:35:32 Magnet Current for MAIN, Imag = 154.9743 A Vtrans = 1.549797 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:35:40 Integrated Voltage Measurement X0 = 0.0300000 m, Delta X = 0.0050000 m VT = 0.002812208 Vs 19:35:47 Integrated Voltage Measurement X0 = 0.0300000 m, Delta X = -0.0050000 m VT = -0.002840207 Vs 19:35:48 Magnet Current for MAIN, Imag = 154.9745 A Vtrans = 1.549798 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:35:56 Integrated Voltage Measurement X0 = 0.0300000 m, Delta X = 0.0050000 m VT = 0.002828858 Vs 19:36:04 Integrated Voltage Measurement X0 = 0.0300000 m, Delta X = -0.0050000 m VT = -0.002834041 Vs 19:36:04 Magnet Current for MAIN, Imag = 154.9742 A Vtrans = 1.549796 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:36:12 Integrated Voltage Measurement X0 = 0.0300000 m, Delta X = 0.0050000 m VT = 0.002818375 Vs 19:36:20 Integrated Voltage Measurement X0 = 0.0300000 m, Delta X = -0.0050000 m VT = -0.002844703 Vs 19:36:20 Integrated Voltage Average and RMS of Measurements X0 = 0.0300000 m, Delta X = 0.0050000 m VT = 0.002831156 +- 0.000003278 Vs 19:36:20 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0300000 m, dx = 0.0050000 m, Nturns = 10.0 VT = 0.002831156 +- 0.000003278 Vs BL = 0.05662313 +- 0.00006556 Tm Remeasure 19:36:20 Integrated Voltage Measurement... 19:36:20 Magnet Current for MAIN, Imag = 154.9742 A Vtrans = 1.549796 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:36:28 Integrated Voltage Measurement X0 = 0.0300000 m, Delta X = 0.0050000 m VT = 0.002820799 Vs 19:36:35 Integrated Voltage Measurement X0 = 0.0300000 m, Delta X = -0.0050000 m VT = -0.002831125 Vs 19:36:36 Magnet Current for MAIN, Imag = 154.9742 A Vtrans = 1.549796 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:36:44 Integrated Voltage Measurement X0 = 0.0300000 m, Delta X = 0.0050000 m VT = 0.002831912 Vs 19:36:51 Integrated Voltage Measurement X0 = 0.0300000 m, Delta X = -0.0050000 m VT = -0.002840696 Vs 19:36:52 Magnet Current for MAIN, Imag = 154.9746 A Vtrans = 1.549799 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:37:00 Integrated Voltage Measurement X0 = 0.0300000 m, Delta X = 0.0050000 m VT = 0.002820423 Vs 19:37:07 Integrated Voltage Measurement X0 = 0.0300000 m, Delta X = -0.0050000 m VT = -0.002840999 Vs 19:37:08 Magnet Current for MAIN, Imag = 154.9746 A Vtrans = 1.549799 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:37:16 Integrated Voltage Measurement X0 = 0.0300000 m, Delta X = 0.0050000 m VT = 0.002825748 Vs 19:37:23 Integrated Voltage Measurement X0 = 0.0300000 m, Delta X = -0.0050000 m VT = -0.002842839 Vs 19:37:23 Integrated Voltage Average and RMS of Measurements X0 = 0.0300000 m, Delta X = 0.0050000 m VT = 0.002831818 +- 0.000003930 Vs 19:37:23 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0300000 m, dx = 0.0050000 m, Nturns = 10.0 VT = 0.002831818 +- 0.000003930 Vs BL = 0.05663635 +- 0.00007859 Tm Remeasure 19:37:23 Integrated Voltage Measurement... 19:37:24 Magnet Current for MAIN, Imag = 154.9741 A Vtrans = 1.549794 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:37:32 Integrated Voltage Measurement X0 = 0.0300000 m, Delta X = 0.0050000 m VT = 0.002792528 Vs 19:37:39 Integrated Voltage Measurement X0 = 0.0300000 m, Delta X = -0.0050000 m VT = -0.002835779 Vs 19:37:40 Magnet Current for MAIN, Imag = 154.9740 A Vtrans = 1.549793 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:37:47 Integrated Voltage Measurement X0 = 0.0300000 m, Delta X = 0.0050000 m VT = 0.002814113 Vs 19:37:55 Integrated Voltage Measurement X0 = 0.0300000 m, Delta X = -0.0050000 m VT = -0.002831180 Vs 19:37:56 Magnet Current for MAIN, Imag = 154.9746 A Vtrans = 1.549799 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:38:03 Integrated Voltage Measurement X0 = 0.0300000 m, Delta X = 0.0050000 m VT = 0.002796241 Vs 19:38:11 Integrated Voltage Measurement X0 = 0.0300000 m, Delta X = -0.0050000 m VT = -0.002831119 Vs 19:38:12 Magnet Current for MAIN, Imag = 154.9746 A Vtrans = 1.549799 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:38:19 Integrated Voltage Measurement X0 = 0.0300000 m, Delta X = 0.0050000 m VT = 0.002836177 Vs 19:38:27 Integrated Voltage Measurement X0 = 0.0300000 m, Delta X = -0.0050000 m VT = -0.002830923 Vs 19:38:27 Integrated Voltage Average and RMS of Measurements X0 = 0.0300000 m, Delta X = 0.0050000 m VT = 0.002821007 +- 0.000008073 Vs 19:38:27 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0300000 m, dx = 0.0050000 m, Nturns = 10.0 VT = 0.002821007 +- 0.000008073 Vs BL = 0.05642015 +- 0.00016145 Tm Remeasure 19:38:27 Integrated Voltage Measurement... 19:38:28 Magnet Current for MAIN, Imag = 154.9743 A Vtrans = 1.549797 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:38:35 Integrated Voltage Measurement X0 = 0.0300000 m, Delta X = 0.0050000 m VT = 0.002848388 Vs 19:38:43 Integrated Voltage Measurement X0 = 0.0300000 m, Delta X = -0.0050000 m VT = -0.002835203 Vs 19:38:44 Magnet Current for MAIN, Imag = 154.9747 A Vtrans = 1.549800 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:38:51 Integrated Voltage Measurement X0 = 0.0300000 m, Delta X = 0.0050000 m VT = 0.002779813 Vs 19:38:59 Integrated Voltage Measurement X0 = 0.0300000 m, Delta X = -0.0050000 m VT = -0.002831676 Vs 19:39:00 Magnet Current for MAIN, Imag = 154.9740 A Vtrans = 1.549793 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:39:07 Integrated Voltage Measurement X0 = 0.0300000 m, Delta X = 0.0050000 m VT = 0.002810656 Vs 19:39:15 Integrated Voltage Measurement X0 = 0.0300000 m, Delta X = -0.0050000 m VT = -0.002835016 Vs 19:39:15 Magnet Current for MAIN, Imag = 154.9746 A Vtrans = 1.549799 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:39:23 Integrated Voltage Measurement X0 = 0.0300000 m, Delta X = 0.0050000 m VT = 0.002823375 Vs 19:39:31 Integrated Voltage Measurement X0 = 0.0300000 m, Delta X = -0.0050000 m VT = -0.002844416 Vs 19:39:31 Integrated Voltage Average and RMS of Measurements X0 = 0.0300000 m, Delta X = 0.0050000 m VT = 0.002826068 +- 0.000013529 Vs 19:39:31 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0300000 m, dx = 0.0050000 m, Nturns = 10.0 VT = 0.002826068 +- 0.000013529 Vs BL = 0.05652136 +- 0.00027057 Tm Remeasure 19:39:31 Integrated Voltage Measurement... 19:39:31 Magnet Current for MAIN, Imag = 154.9746 A Vtrans = 1.549799 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:39:39 Integrated Voltage Measurement X0 = 0.0300000 m, Delta X = 0.0050000 m VT = 0.002831674 Vs 19:39:46 Integrated Voltage Measurement X0 = 0.0300000 m, Delta X = -0.0050000 m VT = -0.002835135 Vs 19:39:47 Magnet Current for MAIN, Imag = 154.9742 A Vtrans = 1.549796 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:39:55 Integrated Voltage Measurement X0 = 0.0300000 m, Delta X = 0.0050000 m VT = 0.002807592 Vs 19:40:02 Integrated Voltage Measurement X0 = 0.0300000 m, Delta X = -0.0050000 m VT = -0.002838466 Vs 19:40:03 Magnet Current for MAIN, Imag = 154.9743 A Vtrans = 1.549797 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:40:11 Integrated Voltage Measurement X0 = 0.0300000 m, Delta X = 0.0050000 m VT = 0.002823004 Vs 19:40:18 Integrated Voltage Measurement X0 = 0.0300000 m, Delta X = -0.0050000 m VT = -0.002838109 Vs 19:40:19 Magnet Current for MAIN, Imag = 154.9738 A Vtrans = 1.549792 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:40:27 Integrated Voltage Measurement X0 = 0.0300000 m, Delta X = 0.0050000 m VT = 0.002794312 Vs 19:40:34 Integrated Voltage Measurement X0 = 0.0300000 m, Delta X = -0.0050000 m VT = -0.002845501 Vs 19:40:34 Integrated Voltage Average and RMS of Measurements X0 = 0.0300000 m, Delta X = 0.0050000 m VT = 0.002826724 +- 0.000005465 Vs 19:40:34 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0300000 m, dx = 0.0050000 m, Nturns = 10.0 VT = 0.002826724 +- 0.000005465 Vs BL = 0.05653448 +- 0.00010929 Tm 19:40:34 Integrated Voltage Measurement... 19:40:35 Magnet Current for MAIN, Imag = 154.9741 A Vtrans = 1.549794 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:40:44 Integrated Voltage Measurement X0 = 0.0350000 m, Delta X = 0.0050000 m VT = 0.002660249 Vs 19:40:51 Integrated Voltage Measurement X0 = 0.0350000 m, Delta X = -0.0050000 m VT = -0.002733410 Vs 19:40:52 Magnet Current for MAIN, Imag = 154.9743 A Vtrans = 1.549797 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:40:59 Integrated Voltage Measurement X0 = 0.0350000 m, Delta X = 0.0050000 m VT = 0.002656674 Vs 19:41:07 Integrated Voltage Measurement X0 = 0.0350000 m, Delta X = -0.0050000 m VT = -0.002721253 Vs 19:41:08 Magnet Current for MAIN, Imag = 154.9741 A Vtrans = 1.549794 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:41:15 Integrated Voltage Measurement X0 = 0.0350000 m, Delta X = 0.0050000 m VT = 0.002663704 Vs 19:41:23 Integrated Voltage Measurement X0 = 0.0350000 m, Delta X = -0.0050000 m VT = -0.002712626 Vs 19:41:24 Magnet Current for MAIN, Imag = 154.9743 A Vtrans = 1.549797 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:41:31 Integrated Voltage Measurement X0 = 0.0350000 m, Delta X = 0.0050000 m VT = 0.002664185 Vs 19:41:39 Integrated Voltage Measurement X0 = 0.0350000 m, Delta X = -0.0050000 m VT = -0.002718037 Vs 19:41:39 Integrated Voltage Average and RMS of Measurements X0 = 0.0350000 m, Delta X = 0.0050000 m VT = 0.002691267 +- 0.000003387 Vs 19:41:39 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0350000 m, dx = 0.0050000 m, Nturns = 10.0 VT = 0.002691267 +- 0.000003387 Vs BL = 0.05382534 +- 0.00006775 Tm Remeasure 19:41:39 Integrated Voltage Measurement... 19:41:40 Magnet Current for MAIN, Imag = 154.9736 A Vtrans = 1.549789 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:41:47 Integrated Voltage Measurement X0 = 0.0350000 m, Delta X = 0.0050000 m VT = 0.002658018 Vs 19:41:55 Integrated Voltage Measurement X0 = 0.0350000 m, Delta X = -0.0050000 m VT = -0.002684664 Vs 19:41:56 Magnet Current for MAIN, Imag = 154.9741 A Vtrans = 1.549794 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:42:03 Integrated Voltage Measurement X0 = 0.0350000 m, Delta X = 0.0050000 m VT = 0.002651780 Vs 19:42:11 Integrated Voltage Measurement X0 = 0.0350000 m, Delta X = -0.0050000 m VT = -0.002682353 Vs 19:42:11 Magnet Current for MAIN, Imag = 154.9741 A Vtrans = 1.549794 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:42:19 Integrated Voltage Measurement X0 = 0.0350000 m, Delta X = 0.0050000 m VT = 0.002653152 Vs 19:42:27 Integrated Voltage Measurement X0 = 0.0350000 m, Delta X = -0.0050000 m VT = -0.002710054 Vs 19:42:27 Magnet Current for MAIN, Imag = 154.9740 A Vtrans = 1.549793 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:42:35 Integrated Voltage Measurement X0 = 0.0350000 m, Delta X = 0.0050000 m VT = 0.002652174 Vs 19:42:43 Integrated Voltage Measurement X0 = 0.0350000 m, Delta X = -0.0050000 m VT = -0.002689207 Vs 19:42:43 Integrated Voltage Average and RMS of Measurements X0 = 0.0350000 m, Delta X = 0.0050000 m VT = 0.002672675 +- 0.000005405 Vs 19:42:43 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0350000 m, dx = 0.0050000 m, Nturns = 10.0 VT = 0.002672675 +- 0.000005405 Vs BL = 0.05345350 +- 0.00010811 Tm Remeasure 19:42:43 Integrated Voltage Measurement... 19:42:43 Magnet Current for MAIN, Imag = 154.9738 A Vtrans = 1.549792 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:42:51 Integrated Voltage Measurement X0 = 0.0350000 m, Delta X = 0.0050000 m VT = 0.002657285 Vs 19:42:59 Integrated Voltage Measurement X0 = 0.0350000 m, Delta X = -0.0050000 m VT = -0.002693312 Vs 19:42:59 Magnet Current for MAIN, Imag = 154.9741 A Vtrans = 1.549794 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:43:07 Integrated Voltage Measurement X0 = 0.0350000 m, Delta X = 0.0050000 m VT = 0.002661678 Vs 19:43:15 Integrated Voltage Measurement X0 = 0.0350000 m, Delta X = -0.0050000 m VT = -0.002665470 Vs 19:43:15 Magnet Current for MAIN, Imag = 154.9741 A Vtrans = 1.549794 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:43:23 Integrated Voltage Measurement X0 = 0.0350000 m, Delta X = 0.0050000 m VT = 0.002653974 Vs 19:43:30 Integrated Voltage Measurement X0 = 0.0350000 m, Delta X = -0.0050000 m VT = -0.002673456 Vs 19:43:31 Magnet Current for MAIN, Imag = 154.9740 A Vtrans = 1.549793 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:43:39 Integrated Voltage Measurement X0 = 0.0350000 m, Delta X = 0.0050000 m VT = 0.002664078 Vs 19:43:46 Integrated Voltage Measurement X0 = 0.0350000 m, Delta X = -0.0050000 m VT = -0.002650407 Vs 19:43:46 Integrated Voltage Average and RMS of Measurements X0 = 0.0350000 m, Delta X = 0.0050000 m VT = 0.002664958 +- 0.000006517 Vs 19:43:46 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0350000 m, dx = 0.0050000 m, Nturns = 10.0 VT = 0.002664958 +- 0.000006517 Vs BL = 0.05329915 +- 0.00013035 Tm Remeasure 19:43:46 Integrated Voltage Measurement... 19:43:47 Magnet Current for MAIN, Imag = 154.9740 A Vtrans = 1.549793 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:43:55 Integrated Voltage Measurement X0 = 0.0350000 m, Delta X = 0.0050000 m VT = 0.002649813 Vs 19:44:02 Integrated Voltage Measurement X0 = 0.0350000 m, Delta X = -0.0050000 m VT = -0.002646637 Vs 19:44:03 Magnet Current for MAIN, Imag = 154.9745 A Vtrans = 1.549798 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:44:11 Integrated Voltage Measurement X0 = 0.0350000 m, Delta X = 0.0050000 m VT = 0.002665507 Vs 19:44:18 Integrated Voltage Measurement X0 = 0.0350000 m, Delta X = -0.0050000 m VT = -0.002719156 Vs 19:44:19 Magnet Current for MAIN, Imag = 154.9740 A Vtrans = 1.549793 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:44:26 Integrated Voltage Measurement X0 = 0.0350000 m, Delta X = 0.0050000 m VT = 0.002665964 Vs 19:44:34 Integrated Voltage Measurement X0 = 0.0350000 m, Delta X = -0.0050000 m VT = -0.002700755 Vs 19:44:35 Magnet Current for MAIN, Imag = 154.9738 A Vtrans = 1.549792 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:44:43 Integrated Voltage Measurement X0 = 0.0350000 m, Delta X = 0.0050000 m VT = 0.002668684 Vs 19:44:50 Integrated Voltage Measurement X0 = 0.0350000 m, Delta X = -0.0050000 m VT = -0.002644489 Vs 19:44:50 Integrated Voltage Average and RMS of Measurements X0 = 0.0350000 m, Delta X = 0.0050000 m VT = 0.002670125 +- 0.000018243 Vs 19:44:50 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0350000 m, dx = 0.0050000 m, Nturns = 10.0 VT = 0.002670125 +- 0.000018243 Vs BL = 0.05340251 +- 0.00036485 Tm Remeasure 19:44:50 Integrated Voltage Measurement... 19:44:51 Magnet Current for MAIN, Imag = 154.9740 A Vtrans = 1.549793 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:44:59 Integrated Voltage Measurement X0 = 0.0350000 m, Delta X = 0.0050000 m VT = 0.002661337 Vs 19:45:06 Integrated Voltage Measurement X0 = 0.0350000 m, Delta X = -0.0050000 m VT = -0.002708671 Vs 19:45:07 Magnet Current for MAIN, Imag = 154.9740 A Vtrans = 1.549793 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:45:15 Integrated Voltage Measurement X0 = 0.0350000 m, Delta X = 0.0050000 m VT = 0.002659734 Vs 19:45:22 Integrated Voltage Measurement X0 = 0.0350000 m, Delta X = -0.0050000 m VT = -0.002633426 Vs 19:45:23 Magnet Current for MAIN, Imag = 154.9740 A Vtrans = 1.549793 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:45:31 Integrated Voltage Measurement X0 = 0.0350000 m, Delta X = 0.0050000 m VT = 0.002659103 Vs 19:45:38 Integrated Voltage Measurement X0 = 0.0350000 m, Delta X = -0.0050000 m VT = -0.002665180 Vs 19:45:39 Magnet Current for MAIN, Imag = 154.9741 A Vtrans = 1.549794 V, Factor V/A = 0.010000 V/A, Zero Current = 0.005346 A 19:45:47 Integrated Voltage Measurement X0 = 0.0350000 m, Delta X = 0.0050000 m VT = 0.002672274 Vs 19:45:54 Integrated Voltage Measurement X0 = 0.0350000 m, Delta X = -0.0050000 m VT = -0.002656400 Vs 19:45:54 Integrated Voltage Average and RMS of Measurements X0 = 0.0350000 m, Delta X = 0.0050000 m VT = 0.002664516 +- 0.000013667 Vs 19:45:54 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0350000 m, dx = 0.0050000 m, Nturns = 10.0 VT = 0.002664516 +- 0.000013667 Vs BL = 0.05329031 +- 0.00027334 Tm 19:45:58 Position moved: 0.000000 m 19:46:01 Position moved: 0.000000 m 19:46:04 Magnet system MAIN current ramping to 0.0000 A...