SLAC Magnetic Measurements Date: 10-21-2009 Time: 15:17:37 Planar Insertion Device Analysis -------------------------------- Data File Header: SLAC Magnetic Measurements Date: 10-21-2009 Time: 15:13:38 Project: LCLS Device Type: Undulator Data Type: Serial Number: L143-112000-15 DATASET Number: 3 Tuning Stage: Fine Tuning Run Number: 10 Measurement Type: Z Scans Operator: YL Comment: find Y at min K Session first run: 10 Measurement Device x: Measurement Device y: SENTRON_441 Data File: s:\magdata\LCLS\Undulator\L143-112000-15\DATASET0003\Fine Tuning\Z Scans\Run 010\bscanz.dat Temperature (C) L143-112000-15:--> T5 19.95 --> T2 19.97 --> T1 19.97 --> T3 19.96 --> Air 20.02 --> T4 19.95 Axes position (m) X:0.026835 Y:0.000351 Initial Hall Probe Zero Measurements: Vx_0i = 0.001422 V Vy_0i = -0.001681 V Final Hall Probe Zero Measurements: Vx_0f = 0.001419 V Vy_0f = -0.001603 V Offset Correction Voff = offset+index*increment index = 0,1...21250 Vx:offset = 0.001422 V, increment = -1.316E-10 V Vy:offset = -0.001681 V, increment = 3.689E-09 V Hall Probe Calibration File: Y: c:\CVI\current_calib\Sentron_441\Sentron_441_y.ini X: c:\cvi\current_calib\HALLx_calib_poly_fit_file.dat Reference Magnet Measurements: Bx_ref = 0.0000xx (T) By_ref = 0.0000xx (T) Magnetic Field Measurements vs Z Analysis Parameters: Input file directory = s:\magdata\LCLS\Undulator\L143-112000-15\DATASET0003\Fine Tuning\Z Scans\Run 010\ Input file name = x+00000_y+100_bscanz.dat Input file type = ZVXYBXYH Correction required = 1 Correction, Bx multiplicative factor = 1.000000 Correction, By multiplicative factor = 1.000000 Correction, Bx additive constant = -0.000021 T Correction, By additive constant = -0.000010 T Correction Bxy additive file = none Correction Bxy additive file type = CORR_ZBXY Electron gamma = 26627.100000 Electron energy = 13.606423 Gev Nominal undulator period = 0.030000 m Nominal K value = 3.485788 Nominal radiation wavelength = 1.496897e-010 m Number of undulator poles = 226 Number of end poles (used in analysis) = 4 Cell length = 3.656000 m Nominal phase change through cell = 709.999940 rad Nominal phase of pole after end poles = 14.137167 rad Save as final result = 0 Result file directory = c:\magdata\slac_und_2\ Result file name = zscan_xn0060yp0002.dat Figures close on exit = 0 Figures visible = 1 !!!!!!!!!!!!!!!!!!!!!!!!!!!!!! The Measured Fields Were Corrected: Bx_corr = CORR_BX_MULT * Bx + CORR_BX_ADD + CORR_BX_ADD_FILE By_corr = CORR_BY_MULT * By + CORR_BY_ADD + CORR_BY_ADD_FILE Parameters: CORR_BX_MULT = 1.000000 CORR_BY_MULT = 1.000000 CORR_BX_ADD = -0.000021 T CORR_BY_ADD = -0.000010 T CORR_BX_ADD_FILE = none CORR_BY_ADD_FILE = none !!!!!!!!!!!!!!!!!!!!!!!!!!!!!! Peak And Zero Field Summary: Number of Poles = 226 Number of End Poles (excluded in statistics) = 4 Peak Fields: Ave = 1.271238 T, RMS = 0.007120 T Peak Field Spacing: Ave = 0.015000 m, RMS = 0.000021 m Zero Field Spacing: Ave = 0.015000 m, RMS = 0.000034 m Average Undulator Period = 0.030000 m Hall Probe Overall Field Integrals: Int Bx dz = -1.3076e-004 Tm Int Int Bx dz dz = -2.2190e-004 Tm^2 Int By dz = 3.7913e-005 Tm Int Int By dz dz = 5.5727e-005 Tm^2 Trajectory Exit Information: X at exit = 1.2278e-006 m X slope at exit = 8.3535e-007 Y at exit = 4.8892e-006 m Y slope at exit = 2.8810e-006 Calculated Undulator Parameters: Keff = 3.486809 Beff = 1.244751 T Lambda_rad = 1.497662e-010 m Calculated K Value At 20 C: (Assumes Keff changes by -.05% per degree C) Measured Average Undulator Temperature = 19.962500 C Measured Keff = 3.486809 Calculated Keff at 20 C = 3.486743 Relative Change (Keff_20C - Keff) / Keff = -0.000019 Phase Fluctuations: Slippage vs Z At Peaks Fitted Slope = 4.99217e-009 Undulator Period = 0.030000 m Radiation Wavelength = 1.49766e-010 m Phase Error (From Slippage vs Z Fit Residuals): Ave = -0.000000 deg, Phase error RMS = 3.185655 deg Phase Relative To Nominal Radiation Wave: Nominal Radiation Wavelength = 1.49690e-010 m Cell Length = 3.656000 m Number Of End Poles = 4 Entrance: Phase Error Of First 10 Poles After The End Poles Ave = 2.009830 deg, RMS = 5.352690 deg Center: Phase Error Of All Poles In Undulator Center: Ave = 11.122217 deg, RMS = 6.353963 deg Exit: Phase Error To Cell Boundary Of Last 10 Poles Before The End Poles Ave = 2.234298 deg, RMS = 3.654190 deg Cell: Phase Change Of Cell = 40704.859 deg Nominal Phase Change Of Cell = 40680.000 deg Phase Error Of Cell = 24.859 deg Calculated Phase Change In The Cell At 20 C: (Relative to the nominal radiation wave) (Assumes Keff changes by -.05% per degree C) Measured Average Undulator Temperature = 19.962500 C Measured Phase Change Of Cell = 40704.859 deg Calculated Phase Change Of Cell At 20 C = 40703.548 deg Phase Change Due To Temperature (Phase_20C - Phase) = -1.311 deg