SLAC Magnetic Measurements Date: 07-12-2017 Time: 11:16:13 Project: LCLS-II Magnet Type: Solenoid Magnet Name: 4051 Serial Number: Run Number: 4 Measurement Device: LS460 Operator: sda Comment: B fields vs z with power supply off. Standardized before measurement Standardization for Main: 2 cycles from -10.0 A to 10.0 A to -10.0 A Test Currents for Main (A): 0.00 11:16:24 Stage Position Measurement, pos = 0.0000000 m 11:16:30 Move Stage, position = 0.000000 11:16:30 MoveZ Devices: Move Device = SXDRIVE, Read Device = SXDRIVE 11:17:22 Move Stage, position = 0.240000 11:18:50 3 Axis Hall Probe Zeroed 11:19:17 3 Axis Hall Probe Type Being Used, type = LS460 11:20:57 Move Stage, position = -0.440000 11:21:00 Magnet Current for Main, Imag = -0.0003 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = 0.000000 A 11:21:01 Magnet Current for Main, Imag = -0.0003 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = 0.000000 A 11:21:02 Magnet Current for Main, Imag = -0.0003 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = 0.000000 A 11:21:02 Magnet Current for Main, Imag = -0.0003 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = 0.000000 A 11:21:02 Magnet Avgerage Current Measurement for Main : Imag = -0.0003 +- 0.0000 A 11:21:03 Initial Zero Current Measurement for Main, Imag_zero = -0.000293625 A 11:21:03 Ambient Measurement Device(s) Being Used, type = None 11:21:03 Magnet Voltage Measurement Device(s) Being Used, type = HP34970 11:21:04 Temperature Measurement Device(s) Being Used, type = HPTHERM_HP34970 11:21:04 XMAG Device(s) Being Used, type = None 11:21:04 Ambient Measurement Device(s) Being Used, type = None 11:21:04 3 Axis Hall Probe Range Set, range = 1 11:21:04 Magnet system Main current ramping to 10.0000 A... 11:21:19 Magnet system Main current ramping to -10.0000 A... 11:21:39 Magnet system Main current ramping to 10.0000 A... 11:22:00 Magnet system Main current ramping to -10.0000 A... 11:22:20 Magnet system Main current ramping to 0.0000 A... 11:22:42 Move Stage, position = -0.440000 11:22:43 Stage Position Measurement, pos = -0.4399550 m 11:22:44 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:22:45 Magnet Current for Main, Imag = -0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:22:45 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:22:46 Magnet Current for Main, Imag = -0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:22:46 Magnet Avgerage Current Measurement for Main : Imag = -0.0000 +- 0.0000 A 11:22:47 Magnet Current for Main, Imag = -0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:22:48 Main Volt Voltage Measurement, V = 0.000002 V 11:22:48 Temperature Measurement, Probe Number = 0, T = 29.581000 C 11:22:52 Temperature Measurement, Probe Number = 1, T = 29.401000 C 11:22:53 Temperature Measurement, Probe Number = 2, T = 27.919000 C 11:22:53 Temperature Measurement, Probe Number = 3, T = 28.415000 C 11:22:54 Temperature Measurement, Probe Number = 4, T = 28.953000 C 11:22:54 Temperature Measurement, Probe Number = 5, T = 24.548000 C 11:22:54 Magnet Test Parameter Measurements P0 = Main, Value = -0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.58 P3 = RTD Outside, Value = 29.40 P4 = Supply Water, Value = 27.92 P5 = Return Water, Value = 28.41 P6 = @ RTD Outside, Value = 28.95 P7 = Ambient, Value = 24.55 11:22:54 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000020, Bz = -0.000030 T 11:23:02 Move Stage, position = -0.438000 11:23:04 Stage Position Measurement, pos = -0.4386100 m 11:23:05 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:23:05 Magnet Current for Main, Imag = -0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:23:06 Magnet Current for Main, Imag = -0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:23:07 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:23:07 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 11:23:07 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:23:08 Main Volt Voltage Measurement, V = 0.000001 V 11:23:09 Temperature Measurement, Probe Number = 0, T = 29.565000 C 11:23:12 Temperature Measurement, Probe Number = 1, T = 29.380000 C 11:23:13 Temperature Measurement, Probe Number = 2, T = 27.938000 C 11:23:13 Temperature Measurement, Probe Number = 3, T = 28.399000 C 11:23:14 Temperature Measurement, Probe Number = 4, T = 28.936000 C 11:23:15 Temperature Measurement, Probe Number = 5, T = 24.713000 C 11:23:15 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.57 P3 = RTD Outside, Value = 29.38 P4 = Supply Water, Value = 27.94 P5 = Return Water, Value = 28.40 P6 = @ RTD Outside, Value = 28.94 P7 = Ambient, Value = 24.71 11:23:15 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000020, Bz = -0.000020 T 11:23:22 Move Stage, position = -0.436000 11:23:24 Stage Position Measurement, pos = -0.4366300 m 11:23:25 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:23:25 Magnet Current for Main, Imag = -0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:23:26 Magnet Current for Main, Imag = -0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:23:27 Magnet Current for Main, Imag = -0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:23:27 Magnet Avgerage Current Measurement for Main : Imag = -0.0000 +- 0.0000 A 11:23:28 Magnet Current for Main, Imag = -0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:23:28 Main Volt Voltage Measurement, V = 0.000001 V 11:23:29 Temperature Measurement, Probe Number = 0, T = 29.550000 C 11:23:32 Temperature Measurement, Probe Number = 1, T = 29.360000 C 11:23:33 Temperature Measurement, Probe Number = 2, T = 27.967000 C 11:23:33 Temperature Measurement, Probe Number = 3, T = 28.405000 C 11:23:34 Temperature Measurement, Probe Number = 4, T = 28.922000 C 11:23:35 Temperature Measurement, Probe Number = 5, T = 24.656000 C 11:23:35 Magnet Test Parameter Measurements P0 = Main, Value = -0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.55 P3 = RTD Outside, Value = 29.36 P4 = Supply Water, Value = 27.97 P5 = Return Water, Value = 28.41 P6 = @ RTD Outside, Value = 28.92 P7 = Ambient, Value = 24.66 11:23:35 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000020, Bz = -0.000020 T 11:23:42 Move Stage, position = -0.434000 11:23:44 Stage Position Measurement, pos = -0.4346450 m 11:23:45 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:23:45 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:23:46 Magnet Current for Main, Imag = -0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:23:47 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:23:47 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 11:23:48 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:23:48 Main Volt Voltage Measurement, V = 0.000002 V 11:23:49 Temperature Measurement, Probe Number = 0, T = 29.539000 C 11:23:52 Temperature Measurement, Probe Number = 1, T = 29.348000 C 11:23:53 Temperature Measurement, Probe Number = 2, T = 27.937000 C 11:23:54 Temperature Measurement, Probe Number = 3, T = 28.402000 C 11:23:54 Temperature Measurement, Probe Number = 4, T = 28.908000 C 11:23:55 Temperature Measurement, Probe Number = 5, T = 24.575000 C 11:23:55 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.54 P3 = RTD Outside, Value = 29.35 P4 = Supply Water, Value = 27.94 P5 = Return Water, Value = 28.40 P6 = @ RTD Outside, Value = 28.91 P7 = Ambient, Value = 24.57 11:23:55 3 Axis Hall Probe Field Measurement, Bx = 0.000010, By = 0.000020, Bz = -0.000020 T 11:24:03 Move Stage, position = -0.432000 11:24:04 Stage Position Measurement, pos = -0.4326400 m 11:24:05 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:24:06 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:24:06 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:24:07 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:24:07 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 11:24:08 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:24:08 Main Volt Voltage Measurement, V = 0.000002 V 11:24:09 Temperature Measurement, Probe Number = 0, T = 29.528000 C 11:24:13 Temperature Measurement, Probe Number = 1, T = 29.332000 C 11:24:13 Temperature Measurement, Probe Number = 2, T = 27.892000 C 11:24:14 Temperature Measurement, Probe Number = 3, T = 28.391000 C 11:24:14 Temperature Measurement, Probe Number = 4, T = 28.898000 C 11:24:15 Temperature Measurement, Probe Number = 5, T = 24.538000 C 11:24:15 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.53 P3 = RTD Outside, Value = 29.33 P4 = Supply Water, Value = 27.89 P5 = Return Water, Value = 28.39 P6 = @ RTD Outside, Value = 28.90 P7 = Ambient, Value = 24.54 11:24:15 3 Axis Hall Probe Field Measurement, Bx = 0.000020, By = 0.000020, Bz = -0.000030 T 11:24:23 Move Stage, position = -0.430000 11:24:24 Stage Position Measurement, pos = -0.4306250 m 11:24:25 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:24:26 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:24:26 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:24:27 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:24:27 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 11:24:28 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:24:28 Main Volt Voltage Measurement, V = 0.000001 V 11:24:29 Temperature Measurement, Probe Number = 0, T = 29.514000 C 11:24:33 Temperature Measurement, Probe Number = 1, T = 29.316000 C 11:24:33 Temperature Measurement, Probe Number = 2, T = 27.899000 C 11:24:34 Temperature Measurement, Probe Number = 3, T = 28.374000 C 11:24:34 Temperature Measurement, Probe Number = 4, T = 28.891000 C 11:24:35 Temperature Measurement, Probe Number = 5, T = 24.511000 C 11:24:35 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.51 P3 = RTD Outside, Value = 29.32 P4 = Supply Water, Value = 27.90 P5 = Return Water, Value = 28.37 P6 = @ RTD Outside, Value = 28.89 P7 = Ambient, Value = 24.51 11:24:35 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000030, Bz = -0.000020 T 11:24:43 Move Stage, position = -0.428000 11:24:44 Stage Position Measurement, pos = -0.4286250 m 11:24:45 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:24:46 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:24:47 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:24:47 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:24:47 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 11:24:48 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:24:49 Main Volt Voltage Measurement, V = 0.000001 V 11:24:49 Temperature Measurement, Probe Number = 0, T = 29.500000 C 11:24:53 Temperature Measurement, Probe Number = 1, T = 29.301000 C 11:24:53 Temperature Measurement, Probe Number = 2, T = 27.917000 C 11:24:54 Temperature Measurement, Probe Number = 3, T = 28.367000 C 11:24:55 Temperature Measurement, Probe Number = 4, T = 28.884000 C 11:24:55 Temperature Measurement, Probe Number = 5, T = 24.451000 C 11:24:55 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.50 P3 = RTD Outside, Value = 29.30 P4 = Supply Water, Value = 27.92 P5 = Return Water, Value = 28.37 P6 = @ RTD Outside, Value = 28.88 P7 = Ambient, Value = 24.45 11:24:55 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000020, Bz = -0.000020 T 11:25:03 Move Stage, position = -0.426000 11:25:05 Stage Position Measurement, pos = -0.4266250 m 11:25:05 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:25:06 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:25:07 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:25:08 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:25:08 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 11:25:08 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:25:09 Main Volt Voltage Measurement, V = 0.000002 V 11:25:09 Temperature Measurement, Probe Number = 0, T = 29.483000 C 11:25:13 Temperature Measurement, Probe Number = 1, T = 29.282000 C 11:25:14 Temperature Measurement, Probe Number = 2, T = 27.930000 C 11:25:14 Temperature Measurement, Probe Number = 3, T = 28.341000 C 11:25:15 Temperature Measurement, Probe Number = 4, T = 28.870000 C 11:25:15 Temperature Measurement, Probe Number = 5, T = 24.408000 C 11:25:15 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.48 P3 = RTD Outside, Value = 29.28 P4 = Supply Water, Value = 27.93 P5 = Return Water, Value = 28.34 P6 = @ RTD Outside, Value = 28.87 P7 = Ambient, Value = 24.41 11:25:15 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000020, Bz = -0.000020 T 11:25:23 Move Stage, position = -0.424000 11:25:25 Stage Position Measurement, pos = -0.4246200 m 11:25:25 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:25:26 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:25:27 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:25:28 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:25:28 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 11:25:28 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:25:29 Main Volt Voltage Measurement, V = 0.000002 V 11:25:30 Temperature Measurement, Probe Number = 0, T = 29.467000 C 11:25:33 Temperature Measurement, Probe Number = 1, T = 29.261000 C 11:25:34 Temperature Measurement, Probe Number = 2, T = 27.934000 C 11:25:34 Temperature Measurement, Probe Number = 3, T = 28.318000 C 11:25:35 Temperature Measurement, Probe Number = 4, T = 28.855000 C 11:25:35 Temperature Measurement, Probe Number = 5, T = 24.361000 C 11:25:35 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.47 P3 = RTD Outside, Value = 29.26 P4 = Supply Water, Value = 27.93 P5 = Return Water, Value = 28.32 P6 = @ RTD Outside, Value = 28.86 P7 = Ambient, Value = 24.36 11:25:36 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000020, Bz = -0.000020 T 11:25:43 Move Stage, position = -0.422000 11:25:45 Stage Position Measurement, pos = -0.4225750 m 11:25:46 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:25:46 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:25:47 Magnet Current for Main, Imag = -0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:25:48 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:25:48 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 11:25:48 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:25:49 Main Volt Voltage Measurement, V = 0.000002 V 11:25:50 Temperature Measurement, Probe Number = 0, T = 29.451000 C 11:25:53 Temperature Measurement, Probe Number = 1, T = 29.240000 C 11:25:54 Temperature Measurement, Probe Number = 2, T = 27.962000 C 11:25:54 Temperature Measurement, Probe Number = 3, T = 28.280000 C 11:25:55 Temperature Measurement, Probe Number = 4, T = 28.839000 C 11:25:56 Temperature Measurement, Probe Number = 5, T = 24.348000 C 11:25:56 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.45 P3 = RTD Outside, Value = 29.24 P4 = Supply Water, Value = 27.96 P5 = Return Water, Value = 28.28 P6 = @ RTD Outside, Value = 28.84 P7 = Ambient, Value = 24.35 11:25:56 3 Axis Hall Probe Field Measurement, Bx = 0.000010, By = 0.000030, Bz = -0.000020 T 11:26:03 Move Stage, position = -0.420000 11:26:05 Stage Position Measurement, pos = -0.4205900 m 11:26:06 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:26:06 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:26:07 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:26:08 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:26:08 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 11:26:09 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:26:09 Main Volt Voltage Measurement, V = 0.000002 V 11:26:10 Temperature Measurement, Probe Number = 0, T = 29.432000 C 11:26:13 Temperature Measurement, Probe Number = 1, T = 29.228000 C 11:26:14 Temperature Measurement, Probe Number = 2, T = 27.967000 C 11:26:14 Temperature Measurement, Probe Number = 3, T = 28.218000 C 11:26:15 Temperature Measurement, Probe Number = 4, T = 28.825000 C 11:26:16 Temperature Measurement, Probe Number = 5, T = 24.206000 C 11:26:16 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.43 P3 = RTD Outside, Value = 29.23 P4 = Supply Water, Value = 27.97 P5 = Return Water, Value = 28.22 P6 = @ RTD Outside, Value = 28.82 P7 = Ambient, Value = 24.21 11:26:16 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000030, Bz = -0.000020 T 11:26:24 Move Stage, position = -0.418000 11:26:25 Stage Position Measurement, pos = -0.4186150 m 11:26:26 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:26:27 Magnet Current for Main, Imag = -0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:26:27 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:26:28 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:26:28 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 11:26:29 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:26:29 Main Volt Voltage Measurement, V = 0.000002 V 11:26:30 Temperature Measurement, Probe Number = 0, T = 29.415000 C 11:26:34 Temperature Measurement, Probe Number = 1, T = 29.206000 C 11:26:34 Temperature Measurement, Probe Number = 2, T = 27.951000 C 11:26:35 Temperature Measurement, Probe Number = 3, T = 28.187000 C 11:26:35 Temperature Measurement, Probe Number = 4, T = 28.811000 C 11:26:36 Temperature Measurement, Probe Number = 5, T = 24.172000 C 11:26:36 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.41 P3 = RTD Outside, Value = 29.21 P4 = Supply Water, Value = 27.95 P5 = Return Water, Value = 28.19 P6 = @ RTD Outside, Value = 28.81 P7 = Ambient, Value = 24.17 11:26:36 3 Axis Hall Probe Field Measurement, Bx = 0.000010, By = 0.000020, Bz = -0.000030 T 11:26:44 Move Stage, position = -0.416000 11:26:45 Stage Position Measurement, pos = -0.4166250 m 11:26:46 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:26:47 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:26:47 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:26:48 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:26:48 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 11:26:49 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:26:49 Main Volt Voltage Measurement, V = 0.000002 V 11:26:50 Temperature Measurement, Probe Number = 0, T = 29.397000 C 11:26:54 Temperature Measurement, Probe Number = 1, T = 29.187000 C 11:26:54 Temperature Measurement, Probe Number = 2, T = 27.924000 C 11:26:55 Temperature Measurement, Probe Number = 3, T = 28.124000 C 11:26:55 Temperature Measurement, Probe Number = 4, T = 28.791000 C 11:26:56 Temperature Measurement, Probe Number = 5, T = 24.026000 C 11:26:56 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.40 P3 = RTD Outside, Value = 29.19 P4 = Supply Water, Value = 27.92 P5 = Return Water, Value = 28.12 P6 = @ RTD Outside, Value = 28.79 P7 = Ambient, Value = 24.03 11:26:56 3 Axis Hall Probe Field Measurement, Bx = 0.000010, By = 0.000030, Bz = -0.000020 T 11:27:04 Move Stage, position = -0.414000 11:27:05 Stage Position Measurement, pos = -0.4146400 m 11:27:06 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:27:07 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:27:08 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:27:08 Magnet Current for Main, Imag = -0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:27:08 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 11:27:09 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:27:10 Main Volt Voltage Measurement, V = 0.000002 V 11:27:10 Temperature Measurement, Probe Number = 0, T = 29.383000 C 11:27:14 Temperature Measurement, Probe Number = 1, T = 29.169000 C 11:27:14 Temperature Measurement, Probe Number = 2, T = 27.897000 C 11:27:15 Temperature Measurement, Probe Number = 3, T = 28.078000 C 11:27:16 Temperature Measurement, Probe Number = 4, T = 28.776000 C 11:27:16 Temperature Measurement, Probe Number = 5, T = 24.058000 C 11:27:16 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.38 P3 = RTD Outside, Value = 29.17 P4 = Supply Water, Value = 27.90 P5 = Return Water, Value = 28.08 P6 = @ RTD Outside, Value = 28.78 P7 = Ambient, Value = 24.06 11:27:16 3 Axis Hall Probe Field Measurement, Bx = 0.000010, By = 0.000020, Bz = -0.000020 T 11:27:24 Move Stage, position = -0.412000 11:27:26 Stage Position Measurement, pos = -0.4126250 m 11:27:26 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:27:27 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:27:28 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:27:28 Magnet Current for Main, Imag = -0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:27:28 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 11:27:29 Magnet Current for Main, Imag = -0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:27:30 Main Volt Voltage Measurement, V = 0.000003 V 11:27:30 Temperature Measurement, Probe Number = 0, T = 29.369000 C 11:27:34 Temperature Measurement, Probe Number = 1, T = 29.148000 C 11:27:35 Temperature Measurement, Probe Number = 2, T = 27.884000 C 11:27:35 Temperature Measurement, Probe Number = 3, T = 28.061000 C 11:27:36 Temperature Measurement, Probe Number = 4, T = 28.754000 C 11:27:36 Temperature Measurement, Probe Number = 5, T = 24.068000 C 11:27:36 Magnet Test Parameter Measurements P0 = Main, Value = -0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.37 P3 = RTD Outside, Value = 29.15 P4 = Supply Water, Value = 27.88 P5 = Return Water, Value = 28.06 P6 = @ RTD Outside, Value = 28.75 P7 = Ambient, Value = 24.07 11:27:36 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000030, Bz = -0.000020 T 11:27:44 Move Stage, position = -0.410000 11:27:46 Stage Position Measurement, pos = -0.4106150 m 11:27:46 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:27:47 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:27:48 Magnet Current for Main, Imag = -0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:27:49 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:27:49 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 11:27:49 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:27:50 Main Volt Voltage Measurement, V = 0.000003 V 11:27:51 Temperature Measurement, Probe Number = 0, T = 29.358000 C 11:27:54 Temperature Measurement, Probe Number = 1, T = 29.139000 C 11:27:55 Temperature Measurement, Probe Number = 2, T = 27.899000 C 11:27:55 Temperature Measurement, Probe Number = 3, T = 28.098000 C 11:27:56 Temperature Measurement, Probe Number = 4, T = 28.738000 C 11:27:57 Temperature Measurement, Probe Number = 5, T = 24.161000 C 11:27:57 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.36 P3 = RTD Outside, Value = 29.14 P4 = Supply Water, Value = 27.90 P5 = Return Water, Value = 28.10 P6 = @ RTD Outside, Value = 28.74 P7 = Ambient, Value = 24.16 11:27:57 3 Axis Hall Probe Field Measurement, Bx = 0.000010, By = 0.000020, Bz = -0.000030 T 11:28:04 Move Stage, position = -0.408000 11:28:06 Stage Position Measurement, pos = -0.4086050 m 11:28:07 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:28:07 Magnet Current for Main, Imag = -0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:28:08 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:28:09 Magnet Current for Main, Imag = -0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:28:09 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 11:28:10 Magnet Current for Main, Imag = -0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:28:10 Main Volt Voltage Measurement, V = 0.000002 V 11:28:11 Temperature Measurement, Probe Number = 0, T = 29.352000 C 11:28:14 Temperature Measurement, Probe Number = 1, T = 29.134000 C 11:28:15 Temperature Measurement, Probe Number = 2, T = 27.922000 C 11:28:16 Temperature Measurement, Probe Number = 3, T = 28.128000 C 11:28:16 Temperature Measurement, Probe Number = 4, T = 28.728000 C 11:28:17 Temperature Measurement, Probe Number = 5, T = 24.155000 C 11:28:17 Magnet Test Parameter Measurements P0 = Main, Value = -0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.35 P3 = RTD Outside, Value = 29.13 P4 = Supply Water, Value = 27.92 P5 = Return Water, Value = 28.13 P6 = @ RTD Outside, Value = 28.73 P7 = Ambient, Value = 24.16 11:28:17 3 Axis Hall Probe Field Measurement, Bx = 0.000010, By = 0.000030, Bz = -0.000020 T 11:28:24 Move Stage, position = -0.406000 11:28:26 Stage Position Measurement, pos = -0.4066100 m 11:28:27 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:28:27 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:28:28 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:28:29 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:28:29 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 11:28:30 Magnet Current for Main, Imag = -0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:28:30 Main Volt Voltage Measurement, V = 0.000003 V 11:28:31 Temperature Measurement, Probe Number = 0, T = 29.345000 C 11:28:34 Temperature Measurement, Probe Number = 1, T = 29.125000 C 11:28:35 Temperature Measurement, Probe Number = 2, T = 27.915000 C 11:28:36 Temperature Measurement, Probe Number = 3, T = 28.120000 C 11:28:36 Temperature Measurement, Probe Number = 4, T = 28.718000 C 11:28:37 Temperature Measurement, Probe Number = 5, T = 24.036000 C 11:28:37 Magnet Test Parameter Measurements P0 = Main, Value = -0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.34 P3 = RTD Outside, Value = 29.12 P4 = Supply Water, Value = 27.91 P5 = Return Water, Value = 28.12 P6 = @ RTD Outside, Value = 28.72 P7 = Ambient, Value = 24.04 11:28:37 3 Axis Hall Probe Field Measurement, Bx = 0.000010, By = 0.000020, Bz = -0.000030 T 11:28:45 Move Stage, position = -0.404000 11:28:46 Stage Position Measurement, pos = -0.4045800 m 11:28:47 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:28:48 Magnet Current for Main, Imag = -0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:28:48 Magnet Current for Main, Imag = -0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:28:49 Magnet Current for Main, Imag = -0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:28:49 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 11:28:50 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:28:50 Main Volt Voltage Measurement, V = 0.000003 V 11:28:51 Temperature Measurement, Probe Number = 0, T = 29.339000 C 11:28:55 Temperature Measurement, Probe Number = 1, T = 29.112000 C 11:28:55 Temperature Measurement, Probe Number = 2, T = 27.874000 C 11:28:56 Temperature Measurement, Probe Number = 3, T = 28.085000 C 11:28:56 Temperature Measurement, Probe Number = 4, T = 28.707000 C 11:28:57 Temperature Measurement, Probe Number = 5, T = 24.023000 C 11:28:57 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.34 P3 = RTD Outside, Value = 29.11 P4 = Supply Water, Value = 27.87 P5 = Return Water, Value = 28.09 P6 = @ RTD Outside, Value = 28.71 P7 = Ambient, Value = 24.02 11:28:57 3 Axis Hall Probe Field Measurement, Bx = 0.000010, By = 0.000030, Bz = -0.000020 T 11:29:05 Move Stage, position = -0.402000 11:29:06 Stage Position Measurement, pos = -0.4025750 m 11:29:07 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:29:08 Magnet Current for Main, Imag = -0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:29:09 Magnet Current for Main, Imag = -0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:29:09 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:29:09 Magnet Avgerage Current Measurement for Main : Imag = -0.0000 +- 0.0000 A 11:29:10 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:29:11 Main Volt Voltage Measurement, V = 0.000003 V 11:29:11 Temperature Measurement, Probe Number = 0, T = 29.332000 C 11:29:15 Temperature Measurement, Probe Number = 1, T = 29.110000 C 11:29:15 Temperature Measurement, Probe Number = 2, T = 27.797000 C 11:29:16 Temperature Measurement, Probe Number = 3, T = 28.093000 C 11:29:17 Temperature Measurement, Probe Number = 4, T = 28.698000 C 11:29:17 Temperature Measurement, Probe Number = 5, T = 24.067000 C 11:29:17 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.33 P3 = RTD Outside, Value = 29.11 P4 = Supply Water, Value = 27.80 P5 = Return Water, Value = 28.09 P6 = @ RTD Outside, Value = 28.70 P7 = Ambient, Value = 24.07 11:29:17 3 Axis Hall Probe Field Measurement, Bx = 0.000010, By = 0.000020, Bz = -0.000010 T 11:29:25 Move Stage, position = -0.400000 11:29:27 Stage Position Measurement, pos = -0.4006000 m 11:29:27 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:29:28 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:29:29 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:29:29 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:29:29 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 11:29:30 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:29:31 Main Volt Voltage Measurement, V = 0.000003 V 11:29:31 Temperature Measurement, Probe Number = 0, T = 29.322000 C 11:29:35 Temperature Measurement, Probe Number = 1, T = 29.100000 C 11:29:35 Temperature Measurement, Probe Number = 2, T = 27.760000 C 11:29:36 Temperature Measurement, Probe Number = 3, T = 28.114000 C 11:29:37 Temperature Measurement, Probe Number = 4, T = 28.692000 C 11:29:37 Temperature Measurement, Probe Number = 5, T = 24.071000 C 11:29:37 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.32 P3 = RTD Outside, Value = 29.10 P4 = Supply Water, Value = 27.76 P5 = Return Water, Value = 28.11 P6 = @ RTD Outside, Value = 28.69 P7 = Ambient, Value = 24.07 11:29:37 3 Axis Hall Probe Field Measurement, Bx = 0.000010, By = 0.000020, Bz = -0.000020 T 11:29:45 Move Stage, position = -0.398000 11:29:47 Stage Position Measurement, pos = -0.3986100 m 11:29:47 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:29:48 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:29:49 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:29:50 Magnet Current for Main, Imag = -0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:29:50 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 11:29:50 Magnet Current for Main, Imag = -0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:29:51 Main Volt Voltage Measurement, V = 0.000003 V 11:29:51 Temperature Measurement, Probe Number = 0, T = 29.314000 C 11:29:55 Temperature Measurement, Probe Number = 1, T = 29.091000 C 11:29:56 Temperature Measurement, Probe Number = 2, T = 27.749000 C 11:29:56 Temperature Measurement, Probe Number = 3, T = 28.140000 C 11:29:57 Temperature Measurement, Probe Number = 4, T = 28.683000 C 11:29:57 Temperature Measurement, Probe Number = 5, T = 24.041000 C 11:29:57 Magnet Test Parameter Measurements P0 = Main, Value = -0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.31 P3 = RTD Outside, Value = 29.09 P4 = Supply Water, Value = 27.75 P5 = Return Water, Value = 28.14 P6 = @ RTD Outside, Value = 28.68 P7 = Ambient, Value = 24.04 11:29:57 3 Axis Hall Probe Field Measurement, Bx = 0.000010, By = 0.000020, Bz = -0.000020 T 11:30:05 Move Stage, position = -0.396000 11:30:07 Stage Position Measurement, pos = -0.3966250 m 11:30:07 Magnet Current for Main, Imag = -0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:30:08 Magnet Current for Main, Imag = -0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:30:09 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:30:10 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:30:10 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 11:30:10 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:30:11 Main Volt Voltage Measurement, V = 0.000003 V 11:30:11 Temperature Measurement, Probe Number = 0, T = 29.311000 C 11:30:15 Temperature Measurement, Probe Number = 1, T = 29.081000 C 11:30:16 Temperature Measurement, Probe Number = 2, T = 27.774000 C 11:30:16 Temperature Measurement, Probe Number = 3, T = 28.181000 C 11:30:17 Temperature Measurement, Probe Number = 4, T = 28.675000 C 11:30:17 Temperature Measurement, Probe Number = 5, T = 24.089000 C 11:30:17 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.31 P3 = RTD Outside, Value = 29.08 P4 = Supply Water, Value = 27.77 P5 = Return Water, Value = 28.18 P6 = @ RTD Outside, Value = 28.68 P7 = Ambient, Value = 24.09 11:30:18 3 Axis Hall Probe Field Measurement, Bx = 0.000010, By = 0.000020, Bz = -0.000020 T 11:30:25 Move Stage, position = -0.394000 11:30:27 Stage Position Measurement, pos = -0.3946200 m 11:30:28 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:30:28 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:30:29 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:30:30 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:30:30 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 11:30:30 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:30:31 Main Volt Voltage Measurement, V = 0.000003 V 11:30:32 Temperature Measurement, Probe Number = 0, T = 29.307000 C 11:30:35 Temperature Measurement, Probe Number = 1, T = 29.078000 C 11:30:36 Temperature Measurement, Probe Number = 2, T = 27.814000 C 11:30:36 Temperature Measurement, Probe Number = 3, T = 28.216000 C 11:30:37 Temperature Measurement, Probe Number = 4, T = 28.670000 C 11:30:38 Temperature Measurement, Probe Number = 5, T = 24.102000 C 11:30:38 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.31 P3 = RTD Outside, Value = 29.08 P4 = Supply Water, Value = 27.81 P5 = Return Water, Value = 28.22 P6 = @ RTD Outside, Value = 28.67 P7 = Ambient, Value = 24.10 11:30:38 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000020, Bz = -0.000030 T 11:30:45 Move Stage, position = -0.392000 11:30:47 Stage Position Measurement, pos = -0.3925950 m 11:30:48 Magnet Current for Main, Imag = -0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:30:48 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:30:49 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:30:50 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:30:50 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 11:30:51 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:30:51 Main Volt Voltage Measurement, V = 0.000002 V 11:30:52 Temperature Measurement, Probe Number = 0, T = 29.306000 C 11:30:55 Temperature Measurement, Probe Number = 1, T = 29.072000 C 11:30:56 Temperature Measurement, Probe Number = 2, T = 27.870000 C 11:30:57 Temperature Measurement, Probe Number = 3, T = 28.255000 C 11:30:57 Temperature Measurement, Probe Number = 4, T = 28.663000 C 11:30:58 Temperature Measurement, Probe Number = 5, T = 24.170000 C 11:30:58 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.31 P3 = RTD Outside, Value = 29.07 P4 = Supply Water, Value = 27.87 P5 = Return Water, Value = 28.25 P6 = @ RTD Outside, Value = 28.66 P7 = Ambient, Value = 24.17 11:30:58 3 Axis Hall Probe Field Measurement, Bx = 0.000010, By = 0.000020, Bz = -0.000020 T 11:31:05 Move Stage, position = -0.390000 11:31:07 Stage Position Measurement, pos = -0.3905850 m 11:31:08 Magnet Current for Main, Imag = -0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:31:09 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:31:09 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:31:10 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:31:10 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 11:31:11 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:31:11 Main Volt Voltage Measurement, V = 0.000003 V 11:31:12 Temperature Measurement, Probe Number = 0, T = 29.309000 C 11:31:15 Temperature Measurement, Probe Number = 1, T = 29.066000 C 11:31:16 Temperature Measurement, Probe Number = 2, T = 27.946000 C 11:31:17 Temperature Measurement, Probe Number = 3, T = 28.291000 C 11:31:17 Temperature Measurement, Probe Number = 4, T = 28.655000 C 11:31:18 Temperature Measurement, Probe Number = 5, T = 24.217000 C 11:31:18 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.31 P3 = RTD Outside, Value = 29.07 P4 = Supply Water, Value = 27.95 P5 = Return Water, Value = 28.29 P6 = @ RTD Outside, Value = 28.66 P7 = Ambient, Value = 24.22 11:31:18 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000030, Bz = -0.000020 T 11:31:26 Move Stage, position = -0.388000 11:31:27 Stage Position Measurement, pos = -0.3885750 m 11:31:28 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:31:29 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:31:29 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:31:30 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:31:30 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 11:31:31 Magnet Current for Main, Imag = -0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:31:31 Main Volt Voltage Measurement, V = 0.000002 V 11:31:32 Temperature Measurement, Probe Number = 0, T = 29.312000 C 11:31:36 Temperature Measurement, Probe Number = 1, T = 29.067000 C 11:31:36 Temperature Measurement, Probe Number = 2, T = 28.001000 C 11:31:37 Temperature Measurement, Probe Number = 3, T = 28.325000 C 11:31:37 Temperature Measurement, Probe Number = 4, T = 28.649000 C 11:31:38 Temperature Measurement, Probe Number = 5, T = 24.261000 C 11:31:38 Magnet Test Parameter Measurements P0 = Main, Value = -0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.31 P3 = RTD Outside, Value = 29.07 P4 = Supply Water, Value = 28.00 P5 = Return Water, Value = 28.32 P6 = @ RTD Outside, Value = 28.65 P7 = Ambient, Value = 24.26 11:31:38 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000020, Bz = -0.000020 T 11:31:46 Move Stage, position = -0.386000 11:31:47 Stage Position Measurement, pos = -0.3865550 m 11:31:48 Magnet Current for Main, Imag = -0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:31:49 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:31:49 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:31:50 Magnet Current for Main, Imag = -0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:31:50 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 11:31:51 Magnet Current for Main, Imag = -0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:31:51 Main Volt Voltage Measurement, V = 0.000003 V 11:31:52 Temperature Measurement, Probe Number = 0, T = 29.313000 C 11:31:56 Temperature Measurement, Probe Number = 1, T = 29.069000 C 11:31:56 Temperature Measurement, Probe Number = 2, T = 28.002000 C 11:31:57 Temperature Measurement, Probe Number = 3, T = 28.338000 C 11:31:57 Temperature Measurement, Probe Number = 4, T = 28.650000 C 11:31:58 Temperature Measurement, Probe Number = 5, T = 24.410000 C 11:31:58 Magnet Test Parameter Measurements P0 = Main, Value = -0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.31 P3 = RTD Outside, Value = 29.07 P4 = Supply Water, Value = 28.00 P5 = Return Water, Value = 28.34 P6 = @ RTD Outside, Value = 28.65 P7 = Ambient, Value = 24.41 11:31:58 3 Axis Hall Probe Field Measurement, Bx = 0.000010, By = 0.000030, Bz = -0.000020 T 11:32:06 Move Stage, position = -0.384000 11:32:07 Stage Position Measurement, pos = -0.3845850 m 11:32:08 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:32:09 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:32:10 Magnet Current for Main, Imag = -0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:32:10 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:32:10 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 11:32:11 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:32:12 Main Volt Voltage Measurement, V = 0.000002 V 11:32:12 Temperature Measurement, Probe Number = 0, T = 29.316000 C 11:32:16 Temperature Measurement, Probe Number = 1, T = 29.072000 C 11:32:16 Temperature Measurement, Probe Number = 2, T = 27.995000 C 11:32:17 Temperature Measurement, Probe Number = 3, T = 28.334000 C 11:32:18 Temperature Measurement, Probe Number = 4, T = 28.650000 C 11:32:18 Temperature Measurement, Probe Number = 5, T = 24.411000 C 11:32:18 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.32 P3 = RTD Outside, Value = 29.07 P4 = Supply Water, Value = 28.00 P5 = Return Water, Value = 28.33 P6 = @ RTD Outside, Value = 28.65 P7 = Ambient, Value = 24.41 11:32:18 3 Axis Hall Probe Field Measurement, Bx = 0.000010, By = 0.000030, Bz = -0.000030 T 11:32:26 Move Stage, position = -0.382000 11:32:28 Stage Position Measurement, pos = -0.3825750 m 11:32:28 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:32:29 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:32:30 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:32:30 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:32:30 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 11:32:31 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:32:32 Main Volt Voltage Measurement, V = 0.000002 V 11:32:32 Temperature Measurement, Probe Number = 0, T = 29.317000 C 11:32:36 Temperature Measurement, Probe Number = 1, T = 29.076000 C 11:32:36 Temperature Measurement, Probe Number = 2, T = 27.978000 C 11:32:37 Temperature Measurement, Probe Number = 3, T = 28.345000 C 11:32:38 Temperature Measurement, Probe Number = 4, T = 28.653000 C 11:32:38 Temperature Measurement, Probe Number = 5, T = 24.461000 C 11:32:38 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.32 P3 = RTD Outside, Value = 29.08 P4 = Supply Water, Value = 27.98 P5 = Return Water, Value = 28.34 P6 = @ RTD Outside, Value = 28.65 P7 = Ambient, Value = 24.46 11:32:38 3 Axis Hall Probe Field Measurement, Bx = 0.000010, By = 0.000030, Bz = -0.000020 T 11:32:46 Move Stage, position = -0.380000 11:32:48 Stage Position Measurement, pos = -0.3805800 m 11:32:48 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:32:49 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:32:50 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:32:51 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:32:51 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 11:32:51 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:32:52 Main Volt Voltage Measurement, V = 0.000002 V 11:32:52 Temperature Measurement, Probe Number = 0, T = 29.318000 C 11:32:56 Temperature Measurement, Probe Number = 1, T = 29.075000 C 11:32:57 Temperature Measurement, Probe Number = 2, T = 27.978000 C 11:32:57 Temperature Measurement, Probe Number = 3, T = 28.359000 C 11:32:58 Temperature Measurement, Probe Number = 4, T = 28.656000 C 11:32:58 Temperature Measurement, Probe Number = 5, T = 24.473000 C 11:32:58 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.32 P3 = RTD Outside, Value = 29.07 P4 = Supply Water, Value = 27.98 P5 = Return Water, Value = 28.36 P6 = @ RTD Outside, Value = 28.66 P7 = Ambient, Value = 24.47 11:32:58 3 Axis Hall Probe Field Measurement, Bx = 0.000010, By = 0.000030, Bz = -0.000020 T 11:33:06 Move Stage, position = -0.378000 11:33:08 Stage Position Measurement, pos = -0.3785950 m 11:33:08 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:33:09 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:33:10 Magnet Current for Main, Imag = -0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:33:11 Magnet Current for Main, Imag = -0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:33:11 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 11:33:11 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:33:12 Main Volt Voltage Measurement, V = 0.000002 V 11:33:12 Temperature Measurement, Probe Number = 0, T = 29.316000 C 11:33:16 Temperature Measurement, Probe Number = 1, T = 29.080000 C 11:33:17 Temperature Measurement, Probe Number = 2, T = 27.949000 C 11:33:17 Temperature Measurement, Probe Number = 3, T = 28.358000 C 11:33:18 Temperature Measurement, Probe Number = 4, T = 28.660000 C 11:33:18 Temperature Measurement, Probe Number = 5, T = 24.453000 C 11:33:18 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.32 P3 = RTD Outside, Value = 29.08 P4 = Supply Water, Value = 27.95 P5 = Return Water, Value = 28.36 P6 = @ RTD Outside, Value = 28.66 P7 = Ambient, Value = 24.45 11:33:19 3 Axis Hall Probe Field Measurement, Bx = 0.000010, By = 0.000040, Bz = -0.000020 T 11:33:26 Move Stage, position = -0.376000 11:33:28 Stage Position Measurement, pos = -0.3766000 m 11:33:29 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:33:29 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:33:30 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:33:31 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:33:31 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 11:33:31 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:33:32 Main Volt Voltage Measurement, V = 0.000002 V 11:33:33 Temperature Measurement, Probe Number = 0, T = 29.309000 C 11:33:36 Temperature Measurement, Probe Number = 1, T = 29.072000 C 11:33:37 Temperature Measurement, Probe Number = 2, T = 27.901000 C 11:33:37 Temperature Measurement, Probe Number = 3, T = 28.359000 C 11:33:38 Temperature Measurement, Probe Number = 4, T = 28.658000 C 11:33:39 Temperature Measurement, Probe Number = 5, T = 24.445000 C 11:33:39 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.31 P3 = RTD Outside, Value = 29.07 P4 = Supply Water, Value = 27.90 P5 = Return Water, Value = 28.36 P6 = @ RTD Outside, Value = 28.66 P7 = Ambient, Value = 24.45 11:33:39 3 Axis Hall Probe Field Measurement, Bx = 0.000010, By = 0.000020, Bz = -0.000010 T 11:33:46 Move Stage, position = -0.374000 11:33:48 Stage Position Measurement, pos = -0.3745900 m 11:33:49 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:33:49 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:33:50 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:33:51 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:33:51 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 11:33:52 Magnet Current for Main, Imag = -0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:33:52 Main Volt Voltage Measurement, V = 0.000002 V 11:33:53 Temperature Measurement, Probe Number = 0, T = 29.308000 C 11:33:56 Temperature Measurement, Probe Number = 1, T = 29.074000 C 11:33:57 Temperature Measurement, Probe Number = 2, T = 27.865000 C 11:33:57 Temperature Measurement, Probe Number = 3, T = 28.353000 C 11:33:58 Temperature Measurement, Probe Number = 4, T = 28.658000 C 11:33:59 Temperature Measurement, Probe Number = 5, T = 24.453000 C 11:33:59 Magnet Test Parameter Measurements P0 = Main, Value = -0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.31 P3 = RTD Outside, Value = 29.07 P4 = Supply Water, Value = 27.86 P5 = Return Water, Value = 28.35 P6 = @ RTD Outside, Value = 28.66 P7 = Ambient, Value = 24.45 11:33:59 3 Axis Hall Probe Field Measurement, Bx = 0.000010, By = 0.000030, Bz = -0.000030 T 11:34:06 Move Stage, position = -0.372000 11:34:08 Stage Position Measurement, pos = -0.3725550 m 11:34:09 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:34:09 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:34:10 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:34:11 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:34:11 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 11:34:12 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:34:12 Main Volt Voltage Measurement, V = 0.000002 V 11:34:13 Temperature Measurement, Probe Number = 0, T = 29.303000 C 11:34:16 Temperature Measurement, Probe Number = 1, T = 29.068000 C 11:34:17 Temperature Measurement, Probe Number = 2, T = 27.814000 C 11:34:18 Temperature Measurement, Probe Number = 3, T = 28.331000 C 11:34:18 Temperature Measurement, Probe Number = 4, T = 28.659000 C 11:34:19 Temperature Measurement, Probe Number = 5, T = 24.391000 C 11:34:19 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.30 P3 = RTD Outside, Value = 29.07 P4 = Supply Water, Value = 27.81 P5 = Return Water, Value = 28.33 P6 = @ RTD Outside, Value = 28.66 P7 = Ambient, Value = 24.39 11:34:19 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000020, Bz = -0.000030 T 11:34:27 Move Stage, position = -0.370000 11:34:28 Stage Position Measurement, pos = -0.3705650 m 11:34:29 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:34:30 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:34:30 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:34:31 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:34:31 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 11:34:32 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:34:32 Main Volt Voltage Measurement, V = 0.000002 V 11:34:33 Temperature Measurement, Probe Number = 0, T = 29.288000 C 11:34:37 Temperature Measurement, Probe Number = 1, T = 29.056000 C 11:34:37 Temperature Measurement, Probe Number = 2, T = 27.764000 C 11:34:38 Temperature Measurement, Probe Number = 3, T = 28.320000 C 11:34:38 Temperature Measurement, Probe Number = 4, T = 28.652000 C 11:34:39 Temperature Measurement, Probe Number = 5, T = 24.369000 C 11:34:39 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.29 P3 = RTD Outside, Value = 29.06 P4 = Supply Water, Value = 27.76 P5 = Return Water, Value = 28.32 P6 = @ RTD Outside, Value = 28.65 P7 = Ambient, Value = 24.37 11:34:39 3 Axis Hall Probe Field Measurement, Bx = 0.000010, By = 0.000020, Bz = -0.000030 T 11:34:47 Move Stage, position = -0.368000 11:34:48 Stage Position Measurement, pos = -0.3685350 m 11:34:49 Magnet Current for Main, Imag = -0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:34:50 Magnet Current for Main, Imag = -0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:34:50 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:34:51 Magnet Current for Main, Imag = -0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:34:51 Magnet Avgerage Current Measurement for Main : Imag = -0.0000 +- 0.0000 A 11:34:52 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:34:52 Main Volt Voltage Measurement, V = 0.000002 V 11:34:53 Temperature Measurement, Probe Number = 0, T = 29.287000 C 11:34:57 Temperature Measurement, Probe Number = 1, T = 29.044000 C 11:34:57 Temperature Measurement, Probe Number = 2, T = 27.779000 C 11:34:58 Temperature Measurement, Probe Number = 3, T = 28.294000 C 11:34:58 Temperature Measurement, Probe Number = 4, T = 28.643000 C 11:34:59 Temperature Measurement, Probe Number = 5, T = 24.322000 C 11:34:59 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.29 P3 = RTD Outside, Value = 29.04 P4 = Supply Water, Value = 27.78 P5 = Return Water, Value = 28.29 P6 = @ RTD Outside, Value = 28.64 P7 = Ambient, Value = 24.32 11:34:59 3 Axis Hall Probe Field Measurement, Bx = 0.000010, By = 0.000020, Bz = -0.000010 T 11:35:07 Move Stage, position = -0.366000 11:35:08 Stage Position Measurement, pos = -0.3665550 m 11:35:09 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:35:10 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:35:11 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:35:11 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:35:11 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 11:35:12 Magnet Current for Main, Imag = -0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:35:13 Main Volt Voltage Measurement, V = 0.000002 V 11:35:13 Temperature Measurement, Probe Number = 0, T = 29.282000 C 11:35:17 Temperature Measurement, Probe Number = 1, T = 29.046000 C 11:35:17 Temperature Measurement, Probe Number = 2, T = 27.825000 C 11:35:18 Temperature Measurement, Probe Number = 3, T = 28.259000 C 11:35:19 Temperature Measurement, Probe Number = 4, T = 28.645000 C 11:35:19 Temperature Measurement, Probe Number = 5, T = 24.376000 C 11:35:19 Magnet Test Parameter Measurements P0 = Main, Value = -0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.28 P3 = RTD Outside, Value = 29.05 P4 = Supply Water, Value = 27.82 P5 = Return Water, Value = 28.26 P6 = @ RTD Outside, Value = 28.64 P7 = Ambient, Value = 24.38 11:35:19 3 Axis Hall Probe Field Measurement, Bx = 0.000010, By = 0.000020, Bz = -0.000030 T 11:35:27 Move Stage, position = -0.364000 11:35:28 Stage Position Measurement, pos = -0.3645600 m 11:35:29 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:35:30 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:35:31 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:35:31 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:35:31 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 11:35:32 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:35:33 Main Volt Voltage Measurement, V = 0.000002 V 11:35:33 Temperature Measurement, Probe Number = 0, T = 29.280000 C 11:35:37 Temperature Measurement, Probe Number = 1, T = 29.045000 C 11:35:38 Temperature Measurement, Probe Number = 2, T = 27.861000 C 11:35:38 Temperature Measurement, Probe Number = 3, T = 28.252000 C 11:35:39 Temperature Measurement, Probe Number = 4, T = 28.645000 C 11:35:39 Temperature Measurement, Probe Number = 5, T = 24.390000 C 11:35:39 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.28 P3 = RTD Outside, Value = 29.05 P4 = Supply Water, Value = 27.86 P5 = Return Water, Value = 28.25 P6 = @ RTD Outside, Value = 28.64 P7 = Ambient, Value = 24.39 11:35:39 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000020, Bz = -0.000010 T 11:35:47 Move Stage, position = -0.362000 11:35:49 Stage Position Measurement, pos = -0.3625550 m 11:35:49 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:35:50 Magnet Current for Main, Imag = -0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:35:51 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:35:52 Magnet Current for Main, Imag = -0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:35:52 Magnet Avgerage Current Measurement for Main : Imag = -0.0000 +- 0.0000 A 11:35:52 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:35:53 Main Volt Voltage Measurement, V = 0.000002 V 11:35:53 Temperature Measurement, Probe Number = 0, T = 29.278000 C 11:35:57 Temperature Measurement, Probe Number = 1, T = 29.040000 C 11:35:58 Temperature Measurement, Probe Number = 2, T = 27.885000 C 11:35:58 Temperature Measurement, Probe Number = 3, T = 28.275000 C 11:35:59 Temperature Measurement, Probe Number = 4, T = 28.644000 C 11:35:59 Temperature Measurement, Probe Number = 5, T = 24.443000 C 11:35:59 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.28 P3 = RTD Outside, Value = 29.04 P4 = Supply Water, Value = 27.89 P5 = Return Water, Value = 28.27 P6 = @ RTD Outside, Value = 28.64 P7 = Ambient, Value = 24.44 11:35:59 3 Axis Hall Probe Field Measurement, Bx = 0.000010, By = 0.000020, Bz = -0.000030 T 11:36:07 Move Stage, position = -0.360000 11:36:09 Stage Position Measurement, pos = -0.3605550 m 11:36:10 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:36:10 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:36:11 Magnet Current for Main, Imag = -0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:36:12 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:36:12 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 11:36:12 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:36:13 Main Volt Voltage Measurement, V = 0.000002 V 11:36:14 Temperature Measurement, Probe Number = 0, T = 29.276000 C 11:36:17 Temperature Measurement, Probe Number = 1, T = 29.036000 C 11:36:18 Temperature Measurement, Probe Number = 2, T = 27.937000 C 11:36:18 Temperature Measurement, Probe Number = 3, T = 28.295000 C 11:36:19 Temperature Measurement, Probe Number = 4, T = 28.643000 C 11:36:20 Temperature Measurement, Probe Number = 5, T = 24.407000 C 11:36:20 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.28 P3 = RTD Outside, Value = 29.04 P4 = Supply Water, Value = 27.94 P5 = Return Water, Value = 28.30 P6 = @ RTD Outside, Value = 28.64 P7 = Ambient, Value = 24.41 11:36:20 3 Axis Hall Probe Field Measurement, Bx = 0.000010, By = 0.000030, Bz = -0.000020 T 11:36:28 Move Stage, position = -0.358000 11:36:29 Stage Position Measurement, pos = -0.3585550 m 11:36:30 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:36:31 Magnet Current for Main, Imag = -0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:36:31 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:36:32 Magnet Current for Main, Imag = -0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:36:32 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 11:36:33 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:36:33 Main Volt Voltage Measurement, V = 0.000002 V 11:36:34 Temperature Measurement, Probe Number = 0, T = 29.275000 C 11:36:38 Temperature Measurement, Probe Number = 1, T = 29.038000 C 11:36:38 Temperature Measurement, Probe Number = 2, T = 27.964000 C 11:36:39 Temperature Measurement, Probe Number = 3, T = 28.266000 C 11:36:39 Temperature Measurement, Probe Number = 4, T = 28.645000 C 11:36:40 Temperature Measurement, Probe Number = 5, T = 24.444000 C 11:36:40 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.27 P3 = RTD Outside, Value = 29.04 P4 = Supply Water, Value = 27.96 P5 = Return Water, Value = 28.27 P6 = @ RTD Outside, Value = 28.64 P7 = Ambient, Value = 24.44 11:36:40 3 Axis Hall Probe Field Measurement, Bx = 0.000010, By = 0.000020, Bz = -0.000020 T 11:36:48 Move Stage, position = -0.356000 11:36:49 Stage Position Measurement, pos = -0.3565450 m 11:36:50 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:36:51 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:36:51 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:36:52 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:36:52 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 11:36:53 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:36:53 Main Volt Voltage Measurement, V = 0.000002 V 11:36:54 Temperature Measurement, Probe Number = 0, T = 29.277000 C 11:36:58 Temperature Measurement, Probe Number = 1, T = 29.034000 C 11:36:58 Temperature Measurement, Probe Number = 2, T = 27.960000 C 11:36:59 Temperature Measurement, Probe Number = 3, T = 28.266000 C 11:36:59 Temperature Measurement, Probe Number = 4, T = 28.641000 C 11:37:00 Temperature Measurement, Probe Number = 5, T = 24.476000 C 11:37:00 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.28 P3 = RTD Outside, Value = 29.03 P4 = Supply Water, Value = 27.96 P5 = Return Water, Value = 28.27 P6 = @ RTD Outside, Value = 28.64 P7 = Ambient, Value = 24.48 11:37:00 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000020, Bz = -0.000030 T 11:37:08 Move Stage, position = -0.354000 11:37:09 Stage Position Measurement, pos = -0.3545150 m 11:37:10 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:37:11 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:37:11 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:37:12 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:37:12 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 11:37:13 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:37:14 Main Volt Voltage Measurement, V = 0.000002 V 11:37:14 Temperature Measurement, Probe Number = 0, T = 29.276000 C 11:37:18 Temperature Measurement, Probe Number = 1, T = 29.035000 C 11:37:18 Temperature Measurement, Probe Number = 2, T = 27.978000 C 11:37:19 Temperature Measurement, Probe Number = 3, T = 28.281000 C 11:37:19 Temperature Measurement, Probe Number = 4, T = 28.641000 C 11:37:20 Temperature Measurement, Probe Number = 5, T = 24.459000 C 11:37:20 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.28 P3 = RTD Outside, Value = 29.04 P4 = Supply Water, Value = 27.98 P5 = Return Water, Value = 28.28 P6 = @ RTD Outside, Value = 28.64 P7 = Ambient, Value = 24.46 11:37:20 3 Axis Hall Probe Field Measurement, Bx = 0.000010, By = 0.000030, Bz = -0.000010 T 11:37:28 Move Stage, position = -0.352000 11:37:29 Stage Position Measurement, pos = -0.3525000 m 11:37:30 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:37:31 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:37:32 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:37:32 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:37:32 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 11:37:33 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:37:34 Main Volt Voltage Measurement, V = 0.000002 V 11:37:34 Temperature Measurement, Probe Number = 0, T = 29.276000 C 11:37:38 Temperature Measurement, Probe Number = 1, T = 29.034000 C 11:37:39 Temperature Measurement, Probe Number = 2, T = 28.008000 C 11:37:39 Temperature Measurement, Probe Number = 3, T = 28.295000 C 11:37:40 Temperature Measurement, Probe Number = 4, T = 28.643000 C 11:37:40 Temperature Measurement, Probe Number = 5, T = 24.447000 C 11:37:40 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.28 P3 = RTD Outside, Value = 29.03 P4 = Supply Water, Value = 28.01 P5 = Return Water, Value = 28.30 P6 = @ RTD Outside, Value = 28.64 P7 = Ambient, Value = 24.45 11:37:40 3 Axis Hall Probe Field Measurement, Bx = 0.000010, By = 0.000030, Bz = -0.000030 T 11:37:48 Move Stage, position = -0.350000 11:37:50 Stage Position Measurement, pos = -0.3504700 m 11:37:50 Magnet Current for Main, Imag = -0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:37:51 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:37:52 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:37:53 Magnet Current for Main, Imag = -0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:37:53 Magnet Avgerage Current Measurement for Main : Imag = -0.0000 +- 0.0000 A 11:37:53 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:37:54 Main Volt Voltage Measurement, V = 0.000002 V 11:37:54 Temperature Measurement, Probe Number = 0, T = 29.274000 C 11:37:58 Temperature Measurement, Probe Number = 1, T = 29.037000 C 11:37:59 Temperature Measurement, Probe Number = 2, T = 28.024000 C 11:37:59 Temperature Measurement, Probe Number = 3, T = 28.299000 C 11:38:00 Temperature Measurement, Probe Number = 4, T = 28.646000 C 11:38:00 Temperature Measurement, Probe Number = 5, T = 24.437000 C 11:38:00 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.27 P3 = RTD Outside, Value = 29.04 P4 = Supply Water, Value = 28.02 P5 = Return Water, Value = 28.30 P6 = @ RTD Outside, Value = 28.65 P7 = Ambient, Value = 24.44 11:38:00 3 Axis Hall Probe Field Measurement, Bx = 0.000010, By = 0.000020, Bz = -0.000030 T 11:38:08 Move Stage, position = -0.348000 11:38:10 Stage Position Measurement, pos = -0.3484900 m 11:38:11 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:38:11 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:38:12 Magnet Current for Main, Imag = -0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:38:13 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:38:13 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 11:38:13 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:38:14 Main Volt Voltage Measurement, V = 0.000002 V 11:38:15 Temperature Measurement, Probe Number = 0, T = 29.273000 C 11:38:18 Temperature Measurement, Probe Number = 1, T = 29.038000 C 11:38:19 Temperature Measurement, Probe Number = 2, T = 28.020000 C 11:38:19 Temperature Measurement, Probe Number = 3, T = 28.315000 C 11:38:20 Temperature Measurement, Probe Number = 4, T = 28.649000 C 11:38:21 Temperature Measurement, Probe Number = 5, T = 24.548000 C 11:38:21 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.27 P3 = RTD Outside, Value = 29.04 P4 = Supply Water, Value = 28.02 P5 = Return Water, Value = 28.32 P6 = @ RTD Outside, Value = 28.65 P7 = Ambient, Value = 24.55 11:38:21 3 Axis Hall Probe Field Measurement, Bx = 0.000010, By = 0.000030, Bz = -0.000020 T 11:38:28 Move Stage, position = -0.346000 11:38:30 Stage Position Measurement, pos = -0.3465000 m 11:38:31 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:38:31 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:38:32 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:38:33 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:38:33 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 11:38:34 Magnet Current for Main, Imag = -0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:38:34 Main Volt Voltage Measurement, V = 0.000002 V 11:38:35 Temperature Measurement, Probe Number = 0, T = 29.274000 C 11:38:38 Temperature Measurement, Probe Number = 1, T = 29.041000 C 11:38:39 Temperature Measurement, Probe Number = 2, T = 27.977000 C 11:38:39 Temperature Measurement, Probe Number = 3, T = 28.324000 C 11:38:40 Temperature Measurement, Probe Number = 4, T = 28.651000 C 11:38:41 Temperature Measurement, Probe Number = 5, T = 24.546000 C 11:38:41 Magnet Test Parameter Measurements P0 = Main, Value = -0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.27 P3 = RTD Outside, Value = 29.04 P4 = Supply Water, Value = 27.98 P5 = Return Water, Value = 28.32 P6 = @ RTD Outside, Value = 28.65 P7 = Ambient, Value = 24.55 11:38:41 3 Axis Hall Probe Field Measurement, Bx = 0.000010, By = 0.000020, Bz = -0.000020 T 11:38:48 Move Stage, position = -0.344000 11:38:50 Stage Position Measurement, pos = -0.3445050 m 11:38:51 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:38:51 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:38:52 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:38:53 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:38:53 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 11:38:54 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:38:54 Main Volt Voltage Measurement, V = 0.000002 V 11:38:55 Temperature Measurement, Probe Number = 0, T = 29.273000 C 11:38:58 Temperature Measurement, Probe Number = 1, T = 29.035000 C 11:38:59 Temperature Measurement, Probe Number = 2, T = 27.937000 C 11:39:00 Temperature Measurement, Probe Number = 3, T = 28.351000 C 11:39:00 Temperature Measurement, Probe Number = 4, T = 28.640000 C 11:39:01 Temperature Measurement, Probe Number = 5, T = 24.571000 C 11:39:01 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.27 P3 = RTD Outside, Value = 29.04 P4 = Supply Water, Value = 27.94 P5 = Return Water, Value = 28.35 P6 = @ RTD Outside, Value = 28.64 P7 = Ambient, Value = 24.57 11:39:01 3 Axis Hall Probe Field Measurement, Bx = 0.000010, By = 0.000020, Bz = -0.000020 T 11:39:09 Move Stage, position = -0.342000 11:39:10 Stage Position Measurement, pos = -0.3424950 m 11:39:11 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:39:12 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:39:12 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:39:13 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:39:13 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 11:39:14 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:39:14 Main Volt Voltage Measurement, V = 0.000002 V 11:39:15 Temperature Measurement, Probe Number = 0, T = 29.277000 C 11:39:19 Temperature Measurement, Probe Number = 1, T = 29.036000 C 11:39:19 Temperature Measurement, Probe Number = 2, T = 27.929000 C 11:39:20 Temperature Measurement, Probe Number = 3, T = 28.380000 C 11:39:20 Temperature Measurement, Probe Number = 4, T = 28.639000 C 11:39:21 Temperature Measurement, Probe Number = 5, T = 24.578000 C 11:39:21 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.28 P3 = RTD Outside, Value = 29.04 P4 = Supply Water, Value = 27.93 P5 = Return Water, Value = 28.38 P6 = @ RTD Outside, Value = 28.64 P7 = Ambient, Value = 24.58 11:39:21 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000020, Bz = -0.000030 T 11:39:29 Move Stage, position = -0.340000 11:39:30 Stage Position Measurement, pos = -0.3405050 m 11:39:31 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:39:32 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:39:32 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:39:33 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:39:33 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 11:39:34 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:39:34 Main Volt Voltage Measurement, V = 0.000002 V 11:39:35 Temperature Measurement, Probe Number = 0, T = 29.282000 C 11:39:39 Temperature Measurement, Probe Number = 1, T = 29.038000 C 11:39:39 Temperature Measurement, Probe Number = 2, T = 27.904000 C 11:39:40 Temperature Measurement, Probe Number = 3, T = 28.383000 C 11:39:40 Temperature Measurement, Probe Number = 4, T = 28.641000 C 11:39:41 Temperature Measurement, Probe Number = 5, T = 24.572000 C 11:39:41 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.28 P3 = RTD Outside, Value = 29.04 P4 = Supply Water, Value = 27.90 P5 = Return Water, Value = 28.38 P6 = @ RTD Outside, Value = 28.64 P7 = Ambient, Value = 24.57 11:39:41 3 Axis Hall Probe Field Measurement, Bx = 0.000010, By = 0.000020, Bz = -0.000010 T 11:39:49 Move Stage, position = -0.338000 11:39:50 Stage Position Measurement, pos = -0.3384850 m 11:39:51 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:39:52 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:39:53 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:39:53 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:39:53 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 11:39:54 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:39:55 Main Volt Voltage Measurement, V = 0.000001 V 11:39:55 Temperature Measurement, Probe Number = 0, T = 29.285000 C 11:39:59 Temperature Measurement, Probe Number = 1, T = 29.041000 C 11:39:59 Temperature Measurement, Probe Number = 2, T = 27.874000 C 11:40:00 Temperature Measurement, Probe Number = 3, T = 28.379000 C 11:40:00 Temperature Measurement, Probe Number = 4, T = 28.645000 C 11:40:01 Temperature Measurement, Probe Number = 5, T = 24.610000 C 11:40:01 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.29 P3 = RTD Outside, Value = 29.04 P4 = Supply Water, Value = 27.87 P5 = Return Water, Value = 28.38 P6 = @ RTD Outside, Value = 28.64 P7 = Ambient, Value = 24.61 11:40:01 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000030, Bz = -0.000020 T 11:40:09 Move Stage, position = -0.336000 11:40:10 Stage Position Measurement, pos = -0.3364750 m 11:40:11 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:40:12 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:40:13 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:40:13 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:40:13 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 11:40:14 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:40:15 Main Volt Voltage Measurement, V = 0.000001 V 11:40:15 Temperature Measurement, Probe Number = 0, T = 29.285000 C 11:40:19 Temperature Measurement, Probe Number = 1, T = 29.043000 C 11:40:19 Temperature Measurement, Probe Number = 2, T = 27.887000 C 11:40:20 Temperature Measurement, Probe Number = 3, T = 28.381000 C 11:40:21 Temperature Measurement, Probe Number = 4, T = 28.648000 C 11:40:21 Temperature Measurement, Probe Number = 5, T = 24.711000 C 11:40:21 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.29 P3 = RTD Outside, Value = 29.04 P4 = Supply Water, Value = 27.89 P5 = Return Water, Value = 28.38 P6 = @ RTD Outside, Value = 28.65 P7 = Ambient, Value = 24.71 11:40:21 3 Axis Hall Probe Field Measurement, Bx = 0.000010, By = 0.000020, Bz = -0.000020 T 11:40:29 Move Stage, position = -0.334000 11:40:31 Stage Position Measurement, pos = -0.3344850 m 11:40:31 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:40:32 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:40:33 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:40:33 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:40:33 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 11:40:34 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:40:35 Main Volt Voltage Measurement, V = 0.000001 V 11:40:35 Temperature Measurement, Probe Number = 0, T = 29.284000 C 11:40:39 Temperature Measurement, Probe Number = 1, T = 29.035000 C 11:40:40 Temperature Measurement, Probe Number = 2, T = 27.932000 C 11:40:40 Temperature Measurement, Probe Number = 3, T = 28.414000 C 11:40:41 Temperature Measurement, Probe Number = 4, T = 28.640000 C 11:40:41 Temperature Measurement, Probe Number = 5, T = 24.772000 C 11:40:41 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.28 P3 = RTD Outside, Value = 29.04 P4 = Supply Water, Value = 27.93 P5 = Return Water, Value = 28.41 P6 = @ RTD Outside, Value = 28.64 P7 = Ambient, Value = 24.77 11:40:41 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000020, Bz = -0.000020 T 11:40:49 Move Stage, position = -0.332000 11:40:51 Stage Position Measurement, pos = -0.3324350 m 11:40:52 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:40:52 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:40:53 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:40:54 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:40:54 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 11:40:54 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:40:55 Main Volt Voltage Measurement, V = 0.000002 V 11:40:56 Temperature Measurement, Probe Number = 0, T = 29.286000 C 11:40:59 Temperature Measurement, Probe Number = 1, T = 29.022000 C 11:41:00 Temperature Measurement, Probe Number = 2, T = 27.939000 C 11:41:00 Temperature Measurement, Probe Number = 3, T = 28.434000 C 11:41:01 Temperature Measurement, Probe Number = 4, T = 28.615000 C 11:41:02 Temperature Measurement, Probe Number = 5, T = 24.728000 C 11:41:02 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.29 P3 = RTD Outside, Value = 29.02 P4 = Supply Water, Value = 27.94 P5 = Return Water, Value = 28.43 P6 = @ RTD Outside, Value = 28.61 P7 = Ambient, Value = 24.73 11:41:02 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000030, Bz = -0.000030 T 11:41:09 Move Stage, position = -0.330000 11:41:11 Stage Position Measurement, pos = -0.3304500 m 11:41:12 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:41:12 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:41:13 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:41:14 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:41:14 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 11:41:15 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:41:15 Main Volt Voltage Measurement, V = 0.000001 V 11:41:16 Temperature Measurement, Probe Number = 0, T = 29.295000 C 11:41:19 Temperature Measurement, Probe Number = 1, T = 29.037000 C 11:41:20 Temperature Measurement, Probe Number = 2, T = 27.946000 C 11:41:20 Temperature Measurement, Probe Number = 3, T = 28.433000 C 11:41:21 Temperature Measurement, Probe Number = 4, T = 28.619000 C 11:41:22 Temperature Measurement, Probe Number = 5, T = 24.754000 C 11:41:22 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.30 P3 = RTD Outside, Value = 29.04 P4 = Supply Water, Value = 27.95 P5 = Return Water, Value = 28.43 P6 = @ RTD Outside, Value = 28.62 P7 = Ambient, Value = 24.75 11:41:22 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000020, Bz = -0.000020 T 11:41:29 Move Stage, position = -0.328000 11:41:31 Stage Position Measurement, pos = -0.3284600 m 11:41:32 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:41:32 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:41:33 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:41:34 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:41:34 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 11:41:35 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:41:35 Main Volt Voltage Measurement, V = 0.000001 V 11:41:36 Temperature Measurement, Probe Number = 0, T = 29.300000 C 11:41:39 Temperature Measurement, Probe Number = 1, T = 29.045000 C 11:41:40 Temperature Measurement, Probe Number = 2, T = 27.975000 C 11:41:41 Temperature Measurement, Probe Number = 3, T = 28.448000 C 11:41:41 Temperature Measurement, Probe Number = 4, T = 28.630000 C 11:41:42 Temperature Measurement, Probe Number = 5, T = 24.773000 C 11:41:42 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.30 P3 = RTD Outside, Value = 29.05 P4 = Supply Water, Value = 27.98 P5 = Return Water, Value = 28.45 P6 = @ RTD Outside, Value = 28.63 P7 = Ambient, Value = 24.77 11:41:42 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000020, Bz = -0.000030 T 11:41:50 Move Stage, position = -0.326000 11:41:51 Stage Position Measurement, pos = -0.3264750 m 11:41:52 Magnet Current for Main, Imag = -0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:41:53 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:41:53 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:41:54 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:41:54 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 11:41:55 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:41:55 Main Volt Voltage Measurement, V = 0.000001 V 11:41:56 Temperature Measurement, Probe Number = 0, T = 29.302000 C 11:42:00 Temperature Measurement, Probe Number = 1, T = 29.053000 C 11:42:00 Temperature Measurement, Probe Number = 2, T = 27.994000 C 11:42:01 Temperature Measurement, Probe Number = 3, T = 28.452000 C 11:42:01 Temperature Measurement, Probe Number = 4, T = 28.641000 C 11:42:02 Temperature Measurement, Probe Number = 5, T = 24.865000 C 11:42:02 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.30 P3 = RTD Outside, Value = 29.05 P4 = Supply Water, Value = 27.99 P5 = Return Water, Value = 28.45 P6 = @ RTD Outside, Value = 28.64 P7 = Ambient, Value = 24.86 11:42:02 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000020, Bz = -0.000020 T 11:42:10 Move Stage, position = -0.324000 11:42:11 Stage Position Measurement, pos = -0.3244600 m 11:42:12 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:42:13 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:42:13 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:42:14 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:42:14 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 11:42:15 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:42:15 Main Volt Voltage Measurement, V = 0.000001 V 11:42:16 Temperature Measurement, Probe Number = 0, T = 29.302000 C 11:42:20 Temperature Measurement, Probe Number = 1, T = 29.060000 C 11:42:20 Temperature Measurement, Probe Number = 2, T = 27.989000 C 11:42:21 Temperature Measurement, Probe Number = 3, T = 28.449000 C 11:42:21 Temperature Measurement, Probe Number = 4, T = 28.650000 C 11:42:22 Temperature Measurement, Probe Number = 5, T = 24.792000 C 11:42:22 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.30 P3 = RTD Outside, Value = 29.06 P4 = Supply Water, Value = 27.99 P5 = Return Water, Value = 28.45 P6 = @ RTD Outside, Value = 28.65 P7 = Ambient, Value = 24.79 11:42:22 3 Axis Hall Probe Field Measurement, Bx = 0.000010, By = 0.000020, Bz = -0.000030 T 11:42:30 Move Stage, position = -0.322000 11:42:31 Stage Position Measurement, pos = -0.3224600 m 11:42:32 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:42:33 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:42:34 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:42:34 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:42:34 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 11:42:35 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:42:36 Main Volt Voltage Measurement, V = 0.000001 V 11:42:36 Temperature Measurement, Probe Number = 0, T = 29.302000 C 11:42:40 Temperature Measurement, Probe Number = 1, T = 29.060000 C 11:42:40 Temperature Measurement, Probe Number = 2, T = 27.990000 C 11:42:41 Temperature Measurement, Probe Number = 3, T = 28.448000 C 11:42:42 Temperature Measurement, Probe Number = 4, T = 28.658000 C 11:42:42 Temperature Measurement, Probe Number = 5, T = 24.893000 C 11:42:42 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.30 P3 = RTD Outside, Value = 29.06 P4 = Supply Water, Value = 27.99 P5 = Return Water, Value = 28.45 P6 = @ RTD Outside, Value = 28.66 P7 = Ambient, Value = 24.89 11:42:42 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000030, Bz = -0.000020 T 11:42:50 Move Stage, position = -0.320000 11:42:51 Stage Position Measurement, pos = -0.3204450 m 11:42:52 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:42:53 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:42:54 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:42:54 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:42:54 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 11:42:55 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:42:56 Main Volt Voltage Measurement, V = 0.000001 V 11:42:56 Temperature Measurement, Probe Number = 0, T = 29.300000 C 11:43:00 Temperature Measurement, Probe Number = 1, T = 29.060000 C 11:43:00 Temperature Measurement, Probe Number = 2, T = 27.992000 C 11:43:01 Temperature Measurement, Probe Number = 3, T = 28.448000 C 11:43:02 Temperature Measurement, Probe Number = 4, T = 28.663000 C 11:43:02 Temperature Measurement, Probe Number = 5, T = 24.815000 C 11:43:02 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.30 P3 = RTD Outside, Value = 29.06 P4 = Supply Water, Value = 27.99 P5 = Return Water, Value = 28.45 P6 = @ RTD Outside, Value = 28.66 P7 = Ambient, Value = 24.82 11:43:02 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000020, Bz = -0.000020 T 11:43:10 Move Stage, position = -0.318000 11:43:12 Stage Position Measurement, pos = -0.3184350 m 11:43:12 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:43:13 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:43:14 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:43:14 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:43:14 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 11:43:15 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:43:16 Main Volt Voltage Measurement, V = 0.000001 V 11:43:16 Temperature Measurement, Probe Number = 0, T = 29.300000 C 11:43:20 Temperature Measurement, Probe Number = 1, T = 29.062000 C 11:43:21 Temperature Measurement, Probe Number = 2, T = 28.020000 C 11:43:21 Temperature Measurement, Probe Number = 3, T = 28.456000 C 11:43:22 Temperature Measurement, Probe Number = 4, T = 28.669000 C 11:43:22 Temperature Measurement, Probe Number = 5, T = 24.761000 C 11:43:22 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.30 P3 = RTD Outside, Value = 29.06 P4 = Supply Water, Value = 28.02 P5 = Return Water, Value = 28.46 P6 = @ RTD Outside, Value = 28.67 P7 = Ambient, Value = 24.76 11:43:22 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000020, Bz = -0.000020 T 11:43:30 Move Stage, position = -0.316000 11:43:32 Stage Position Measurement, pos = -0.3164300 m 11:43:32 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:43:33 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:43:34 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:43:35 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:43:35 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 11:43:35 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:43:36 Main Volt Voltage Measurement, V = 0.000001 V 11:43:36 Temperature Measurement, Probe Number = 0, T = 29.301000 C 11:43:40 Temperature Measurement, Probe Number = 1, T = 29.059000 C 11:43:41 Temperature Measurement, Probe Number = 2, T = 28.059000 C 11:43:41 Temperature Measurement, Probe Number = 3, T = 28.468000 C 11:43:42 Temperature Measurement, Probe Number = 4, T = 28.670000 C 11:43:42 Temperature Measurement, Probe Number = 5, T = 24.779000 C 11:43:42 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.30 P3 = RTD Outside, Value = 29.06 P4 = Supply Water, Value = 28.06 P5 = Return Water, Value = 28.47 P6 = @ RTD Outside, Value = 28.67 P7 = Ambient, Value = 24.78 11:43:43 3 Axis Hall Probe Field Measurement, Bx = 0.000010, By = 0.000020, Bz = -0.000020 T 11:43:50 Move Stage, position = -0.314000 11:43:52 Stage Position Measurement, pos = -0.3144050 m 11:43:53 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:43:53 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:43:54 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:43:55 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:43:55 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 11:43:55 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:43:56 Main Volt Voltage Measurement, V = 0.000002 V 11:43:57 Temperature Measurement, Probe Number = 0, T = 29.299000 C 11:44:00 Temperature Measurement, Probe Number = 1, T = 29.058000 C 11:44:01 Temperature Measurement, Probe Number = 2, T = 28.096000 C 11:44:01 Temperature Measurement, Probe Number = 3, T = 28.482000 C 11:44:02 Temperature Measurement, Probe Number = 4, T = 28.672000 C 11:44:03 Temperature Measurement, Probe Number = 5, T = 24.801000 C 11:44:03 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.30 P3 = RTD Outside, Value = 29.06 P4 = Supply Water, Value = 28.10 P5 = Return Water, Value = 28.48 P6 = @ RTD Outside, Value = 28.67 P7 = Ambient, Value = 24.80 11:44:03 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000030, Bz = -0.000030 T 11:44:10 Move Stage, position = -0.312000 11:44:12 Stage Position Measurement, pos = -0.3124150 m 11:44:13 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:44:13 Magnet Current for Main, Imag = -0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:44:14 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:44:15 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:44:15 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 11:44:16 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:44:16 Main Volt Voltage Measurement, V = 0.000002 V 11:44:17 Temperature Measurement, Probe Number = 0, T = 29.300000 C 11:44:20 Temperature Measurement, Probe Number = 1, T = 29.058000 C 11:44:21 Temperature Measurement, Probe Number = 2, T = 28.137000 C 11:44:22 Temperature Measurement, Probe Number = 3, T = 28.496000 C 11:44:22 Temperature Measurement, Probe Number = 4, T = 28.674000 C 11:44:23 Temperature Measurement, Probe Number = 5, T = 24.794000 C 11:44:23 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.30 P3 = RTD Outside, Value = 29.06 P4 = Supply Water, Value = 28.14 P5 = Return Water, Value = 28.50 P6 = @ RTD Outside, Value = 28.67 P7 = Ambient, Value = 24.79 11:44:23 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000030, Bz = -0.000020 T 11:44:30 Move Stage, position = -0.310000 11:44:32 Stage Position Measurement, pos = -0.3104150 m 11:44:33 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:44:34 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:44:34 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:44:35 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:44:35 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 11:44:36 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:44:36 Main Volt Voltage Measurement, V = 0.000001 V 11:44:37 Temperature Measurement, Probe Number = 0, T = 29.303000 C 11:44:41 Temperature Measurement, Probe Number = 1, T = 29.052000 C 11:44:41 Temperature Measurement, Probe Number = 2, T = 28.185000 C 11:44:42 Temperature Measurement, Probe Number = 3, T = 28.493000 C 11:44:42 Temperature Measurement, Probe Number = 4, T = 28.672000 C 11:44:43 Temperature Measurement, Probe Number = 5, T = 24.773000 C 11:44:43 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.30 P3 = RTD Outside, Value = 29.05 P4 = Supply Water, Value = 28.18 P5 = Return Water, Value = 28.49 P6 = @ RTD Outside, Value = 28.67 P7 = Ambient, Value = 24.77 11:44:43 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000020, Bz = -0.000020 T 11:44:51 Move Stage, position = -0.308000 11:44:52 Stage Position Measurement, pos = -0.3084450 m 11:44:53 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:44:54 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:44:54 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:44:55 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:44:55 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 11:44:56 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:44:56 Main Volt Voltage Measurement, V = 0.000002 V 11:44:57 Temperature Measurement, Probe Number = 0, T = 29.303000 C 11:45:01 Temperature Measurement, Probe Number = 1, T = 29.054000 C 11:45:01 Temperature Measurement, Probe Number = 2, T = 28.205000 C 11:45:02 Temperature Measurement, Probe Number = 3, T = 28.489000 C 11:45:02 Temperature Measurement, Probe Number = 4, T = 28.673000 C 11:45:03 Temperature Measurement, Probe Number = 5, T = 24.752000 C 11:45:03 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.30 P3 = RTD Outside, Value = 29.05 P4 = Supply Water, Value = 28.20 P5 = Return Water, Value = 28.49 P6 = @ RTD Outside, Value = 28.67 P7 = Ambient, Value = 24.75 11:45:03 3 Axis Hall Probe Field Measurement, Bx = 0.000010, By = 0.000020, Bz = -0.000020 T 11:45:11 Move Stage, position = -0.306000 11:45:12 Stage Position Measurement, pos = -0.3064550 m 11:45:13 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:45:14 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:45:14 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:45:15 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:45:15 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 11:45:16 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:45:16 Main Volt Voltage Measurement, V = 0.000001 V 11:45:17 Temperature Measurement, Probe Number = 0, T = 29.303000 C 11:45:21 Temperature Measurement, Probe Number = 1, T = 29.058000 C 11:45:21 Temperature Measurement, Probe Number = 2, T = 28.203000 C 11:45:22 Temperature Measurement, Probe Number = 3, T = 28.498000 C 11:45:22 Temperature Measurement, Probe Number = 4, T = 28.677000 C 11:45:23 Temperature Measurement, Probe Number = 5, T = 24.754000 C 11:45:23 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.30 P3 = RTD Outside, Value = 29.06 P4 = Supply Water, Value = 28.20 P5 = Return Water, Value = 28.50 P6 = @ RTD Outside, Value = 28.68 P7 = Ambient, Value = 24.75 11:45:23 3 Axis Hall Probe Field Measurement, Bx = 0.000010, By = 0.000030, Bz = -0.000020 T 11:45:31 Move Stage, position = -0.304000 11:45:32 Stage Position Measurement, pos = -0.3044550 m 11:45:33 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:45:34 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:45:35 Magnet Current for Main, Imag = -0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:45:35 Magnet Current for Main, Imag = -0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:45:35 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 11:45:36 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:45:37 Main Volt Voltage Measurement, V = 0.000002 V 11:45:37 Temperature Measurement, Probe Number = 0, T = 29.299000 C 11:45:41 Temperature Measurement, Probe Number = 1, T = 29.054000 C 11:45:41 Temperature Measurement, Probe Number = 2, T = 28.217000 C 11:45:42 Temperature Measurement, Probe Number = 3, T = 28.472000 C 11:45:43 Temperature Measurement, Probe Number = 4, T = 28.679000 C 11:45:43 Temperature Measurement, Probe Number = 5, T = 24.769000 C 11:45:43 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.30 P3 = RTD Outside, Value = 29.05 P4 = Supply Water, Value = 28.22 P5 = Return Water, Value = 28.47 P6 = @ RTD Outside, Value = 28.68 P7 = Ambient, Value = 24.77 11:45:43 3 Axis Hall Probe Field Measurement, Bx = 0.000010, By = 0.000020, Bz = -0.000020 T 11:45:51 Move Stage, position = -0.302000 11:45:53 Stage Position Measurement, pos = -0.3024450 m 11:45:53 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:45:54 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:45:55 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:45:55 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:45:55 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 11:45:56 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:45:57 Main Volt Voltage Measurement, V = 0.000001 V 11:45:57 Temperature Measurement, Probe Number = 0, T = 29.296000 C 11:46:01 Temperature Measurement, Probe Number = 1, T = 29.052000 C 11:46:02 Temperature Measurement, Probe Number = 2, T = 28.234000 C 11:46:02 Temperature Measurement, Probe Number = 3, T = 28.443000 C 11:46:03 Temperature Measurement, Probe Number = 4, T = 28.673000 C 11:46:03 Temperature Measurement, Probe Number = 5, T = 24.970000 C 11:46:03 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.30 P3 = RTD Outside, Value = 29.05 P4 = Supply Water, Value = 28.23 P5 = Return Water, Value = 28.44 P6 = @ RTD Outside, Value = 28.67 P7 = Ambient, Value = 24.97 11:46:03 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000040, Bz = -0.000030 T 11:46:11 Move Stage, position = -0.300000 11:46:13 Stage Position Measurement, pos = -0.3004250 m 11:46:13 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:46:14 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:46:15 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:46:16 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:46:16 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 11:46:16 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:46:17 Main Volt Voltage Measurement, V = 0.000001 V 11:46:17 Temperature Measurement, Probe Number = 0, T = 29.294000 C 11:46:21 Temperature Measurement, Probe Number = 1, T = 29.049000 C 11:46:22 Temperature Measurement, Probe Number = 2, T = 28.269000 C 11:46:22 Temperature Measurement, Probe Number = 3, T = 28.438000 C 11:46:23 Temperature Measurement, Probe Number = 4, T = 28.667000 C 11:46:23 Temperature Measurement, Probe Number = 5, T = 25.098000 C 11:46:23 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.29 P3 = RTD Outside, Value = 29.05 P4 = Supply Water, Value = 28.27 P5 = Return Water, Value = 28.44 P6 = @ RTD Outside, Value = 28.67 P7 = Ambient, Value = 25.10 11:46:23 3 Axis Hall Probe Field Measurement, Bx = -0.000010, By = 0.000030, Bz = -0.000030 T 11:46:31 Move Stage, position = -0.298000 11:46:33 Stage Position Measurement, pos = -0.2984250 m 11:46:33 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:46:34 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:46:35 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:46:36 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:46:36 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 11:46:36 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:46:37 Main Volt Voltage Measurement, V = 0.000001 V 11:46:38 Temperature Measurement, Probe Number = 0, T = 29.298000 C 11:46:41 Temperature Measurement, Probe Number = 1, T = 29.053000 C 11:46:42 Temperature Measurement, Probe Number = 2, T = 28.287000 C 11:46:42 Temperature Measurement, Probe Number = 3, T = 28.467000 C 11:46:43 Temperature Measurement, Probe Number = 4, T = 28.664000 C 11:46:43 Temperature Measurement, Probe Number = 5, T = 24.971000 C 11:46:43 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.30 P3 = RTD Outside, Value = 29.05 P4 = Supply Water, Value = 28.29 P5 = Return Water, Value = 28.47 P6 = @ RTD Outside, Value = 28.66 P7 = Ambient, Value = 24.97 11:46:44 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000020, Bz = -0.000030 T 11:46:51 Move Stage, position = -0.296000 11:46:53 Stage Position Measurement, pos = -0.2964050 m 11:46:54 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:46:54 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:46:55 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:46:56 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:46:56 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 11:46:56 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:46:57 Main Volt Voltage Measurement, V = 0.000001 V 11:46:58 Temperature Measurement, Probe Number = 0, T = 29.304000 C 11:47:01 Temperature Measurement, Probe Number = 1, T = 29.057000 C 11:47:02 Temperature Measurement, Probe Number = 2, T = 28.253000 C 11:47:02 Temperature Measurement, Probe Number = 3, T = 28.477000 C 11:47:03 Temperature Measurement, Probe Number = 4, T = 28.669000 C 11:47:04 Temperature Measurement, Probe Number = 5, T = 24.959000 C 11:47:04 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.30 P3 = RTD Outside, Value = 29.06 P4 = Supply Water, Value = 28.25 P5 = Return Water, Value = 28.48 P6 = @ RTD Outside, Value = 28.67 P7 = Ambient, Value = 24.96 11:47:04 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000030, Bz = -0.000020 T 11:47:11 Move Stage, position = -0.294000 11:47:13 Stage Position Measurement, pos = -0.2944300 m 11:47:14 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:47:14 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:47:15 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:47:16 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:47:16 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 11:47:17 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:47:17 Main Volt Voltage Measurement, V = 0.000001 V 11:47:18 Temperature Measurement, Probe Number = 0, T = 29.305000 C 11:47:21 Temperature Measurement, Probe Number = 1, T = 29.063000 C 11:47:22 Temperature Measurement, Probe Number = 2, T = 28.195000 C 11:47:23 Temperature Measurement, Probe Number = 3, T = 28.467000 C 11:47:23 Temperature Measurement, Probe Number = 4, T = 28.677000 C 11:47:24 Temperature Measurement, Probe Number = 5, T = 24.892000 C 11:47:24 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.30 P3 = RTD Outside, Value = 29.06 P4 = Supply Water, Value = 28.20 P5 = Return Water, Value = 28.47 P6 = @ RTD Outside, Value = 28.68 P7 = Ambient, Value = 24.89 11:47:24 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000030, Bz = -0.000020 T 11:47:31 Move Stage, position = -0.292000 11:47:33 Stage Position Measurement, pos = -0.2924200 m 11:47:34 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:47:35 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:47:35 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:47:36 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:47:36 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 11:47:37 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:47:37 Main Volt Voltage Measurement, V = 0.000001 V 11:47:38 Temperature Measurement, Probe Number = 0, T = 29.299000 C 11:47:42 Temperature Measurement, Probe Number = 1, T = 29.062000 C 11:47:42 Temperature Measurement, Probe Number = 2, T = 28.170000 C 11:47:43 Temperature Measurement, Probe Number = 3, T = 28.443000 C 11:47:43 Temperature Measurement, Probe Number = 4, T = 28.683000 C 11:47:44 Temperature Measurement, Probe Number = 5, T = 24.798000 C 11:47:44 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.30 P3 = RTD Outside, Value = 29.06 P4 = Supply Water, Value = 28.17 P5 = Return Water, Value = 28.44 P6 = @ RTD Outside, Value = 28.68 P7 = Ambient, Value = 24.80 11:47:44 3 Axis Hall Probe Field Measurement, Bx = 0.000010, By = 0.000020, Bz = -0.000020 T 11:47:52 Move Stage, position = -0.290000 11:47:53 Stage Position Measurement, pos = -0.2904350 m 11:47:54 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:47:55 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:47:55 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:47:56 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:47:56 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 11:47:57 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:47:58 Main Volt Voltage Measurement, V = 0.000001 V 11:47:58 Temperature Measurement, Probe Number = 0, T = 29.296000 C 11:48:02 Temperature Measurement, Probe Number = 1, T = 29.057000 C 11:48:02 Temperature Measurement, Probe Number = 2, T = 28.145000 C 11:48:03 Temperature Measurement, Probe Number = 3, T = 28.430000 C 11:48:03 Temperature Measurement, Probe Number = 4, T = 28.683000 C 11:48:04 Temperature Measurement, Probe Number = 5, T = 24.821000 C 11:48:04 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.30 P3 = RTD Outside, Value = 29.06 P4 = Supply Water, Value = 28.14 P5 = Return Water, Value = 28.43 P6 = @ RTD Outside, Value = 28.68 P7 = Ambient, Value = 24.82 11:48:04 3 Axis Hall Probe Field Measurement, Bx = 0.000010, By = 0.000020, Bz = -0.000030 T 11:48:12 Move Stage, position = -0.288000 11:48:13 Stage Position Measurement, pos = -0.2884450 m 11:48:14 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:48:15 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:48:16 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:48:16 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:48:16 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 11:48:17 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:48:18 Main Volt Voltage Measurement, V = 0.000002 V 11:48:18 Temperature Measurement, Probe Number = 0, T = 29.292000 C 11:48:22 Temperature Measurement, Probe Number = 1, T = 29.053000 C 11:48:22 Temperature Measurement, Probe Number = 2, T = 28.136000 C 11:48:23 Temperature Measurement, Probe Number = 3, T = 28.422000 C 11:48:24 Temperature Measurement, Probe Number = 4, T = 28.686000 C 11:48:24 Temperature Measurement, Probe Number = 5, T = 24.841000 C 11:48:24 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.29 P3 = RTD Outside, Value = 29.05 P4 = Supply Water, Value = 28.14 P5 = Return Water, Value = 28.42 P6 = @ RTD Outside, Value = 28.69 P7 = Ambient, Value = 24.84 11:48:24 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000020, Bz = -0.000030 T 11:48:32 Move Stage, position = -0.286000 11:48:34 Stage Position Measurement, pos = -0.2864550 m 11:48:34 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:48:35 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:48:36 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:48:36 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:48:36 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 11:48:37 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:48:38 Main Volt Voltage Measurement, V = 0.000001 V 11:48:38 Temperature Measurement, Probe Number = 0, T = 29.289000 C 11:48:42 Temperature Measurement, Probe Number = 1, T = 29.049000 C 11:48:43 Temperature Measurement, Probe Number = 2, T = 28.130000 C 11:48:43 Temperature Measurement, Probe Number = 3, T = 28.416000 C 11:48:44 Temperature Measurement, Probe Number = 4, T = 28.687000 C 11:48:44 Temperature Measurement, Probe Number = 5, T = 24.797000 C 11:48:44 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.29 P3 = RTD Outside, Value = 29.05 P4 = Supply Water, Value = 28.13 P5 = Return Water, Value = 28.42 P6 = @ RTD Outside, Value = 28.69 P7 = Ambient, Value = 24.80 11:48:44 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000030, Bz = -0.000020 T 11:48:52 Move Stage, position = -0.284000 11:48:54 Stage Position Measurement, pos = -0.2844600 m 11:48:54 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:48:55 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:48:56 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:48:57 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:48:57 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 11:48:57 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:48:58 Main Volt Voltage Measurement, V = 0.000001 V 11:48:58 Temperature Measurement, Probe Number = 0, T = 29.288000 C 11:49:02 Temperature Measurement, Probe Number = 1, T = 29.046000 C 11:49:03 Temperature Measurement, Probe Number = 2, T = 28.124000 C 11:49:03 Temperature Measurement, Probe Number = 3, T = 28.412000 C 11:49:04 Temperature Measurement, Probe Number = 4, T = 28.683000 C 11:49:04 Temperature Measurement, Probe Number = 5, T = 24.642000 C 11:49:04 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.29 P3 = RTD Outside, Value = 29.05 P4 = Supply Water, Value = 28.12 P5 = Return Water, Value = 28.41 P6 = @ RTD Outside, Value = 28.68 P7 = Ambient, Value = 24.64 11:49:04 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000030, Bz = -0.000020 T 11:49:12 Move Stage, position = -0.282000 11:49:14 Stage Position Measurement, pos = -0.2824200 m 11:49:15 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:49:15 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:49:16 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:49:17 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:49:17 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 11:49:17 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:49:18 Main Volt Voltage Measurement, V = 0.000002 V 11:49:19 Temperature Measurement, Probe Number = 0, T = 29.274000 C 11:49:22 Temperature Measurement, Probe Number = 1, T = 29.031000 C 11:49:23 Temperature Measurement, Probe Number = 2, T = 28.056000 C 11:49:23 Temperature Measurement, Probe Number = 3, T = 28.342000 C 11:49:24 Temperature Measurement, Probe Number = 4, T = 28.672000 C 11:49:25 Temperature Measurement, Probe Number = 5, T = 24.559000 C 11:49:25 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.27 P3 = RTD Outside, Value = 29.03 P4 = Supply Water, Value = 28.06 P5 = Return Water, Value = 28.34 P6 = @ RTD Outside, Value = 28.67 P7 = Ambient, Value = 24.56 11:49:25 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000030, Bz = -0.000020 T 11:49:32 Move Stage, position = -0.280000 11:49:34 Stage Position Measurement, pos = -0.2804300 m 11:49:35 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:49:35 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:49:36 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:49:37 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:49:37 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 11:49:37 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:49:38 Main Volt Voltage Measurement, V = 0.000002 V 11:49:39 Temperature Measurement, Probe Number = 0, T = 29.276000 C 11:49:42 Temperature Measurement, Probe Number = 1, T = 29.032000 C 11:49:43 Temperature Measurement, Probe Number = 2, T = 28.024000 C 11:49:43 Temperature Measurement, Probe Number = 3, T = 28.276000 C 11:49:44 Temperature Measurement, Probe Number = 4, T = 28.665000 C 11:49:45 Temperature Measurement, Probe Number = 5, T = 24.565000 C 11:49:45 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.28 P3 = RTD Outside, Value = 29.03 P4 = Supply Water, Value = 28.02 P5 = Return Water, Value = 28.28 P6 = @ RTD Outside, Value = 28.66 P7 = Ambient, Value = 24.57 11:49:45 3 Axis Hall Probe Field Measurement, Bx = 0.000010, By = 0.000030, Bz = -0.000010 T 11:49:52 Move Stage, position = -0.278000 11:49:54 Stage Position Measurement, pos = -0.2784150 m 11:49:55 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:49:55 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:49:56 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:49:57 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:49:57 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 11:49:58 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:49:58 Main Volt Voltage Measurement, V = 0.000002 V 11:49:59 Temperature Measurement, Probe Number = 0, T = 29.275000 C 11:50:02 Temperature Measurement, Probe Number = 1, T = 29.029000 C 11:50:03 Temperature Measurement, Probe Number = 2, T = 28.001000 C 11:50:04 Temperature Measurement, Probe Number = 3, T = 28.256000 C 11:50:04 Temperature Measurement, Probe Number = 4, T = 28.664000 C 11:50:05 Temperature Measurement, Probe Number = 5, T = 24.516000 C 11:50:05 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.27 P3 = RTD Outside, Value = 29.03 P4 = Supply Water, Value = 28.00 P5 = Return Water, Value = 28.26 P6 = @ RTD Outside, Value = 28.66 P7 = Ambient, Value = 24.52 11:50:05 3 Axis Hall Probe Field Measurement, Bx = 0.000010, By = 0.000020, Bz = -0.000040 T 11:50:13 Move Stage, position = -0.276000 11:50:14 Stage Position Measurement, pos = -0.2764400 m 11:50:15 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:50:16 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:50:16 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:50:17 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:50:17 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 11:50:18 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:50:18 Main Volt Voltage Measurement, V = 0.000002 V 11:50:19 Temperature Measurement, Probe Number = 0, T = 29.269000 C 11:50:23 Temperature Measurement, Probe Number = 1, T = 29.022000 C 11:50:23 Temperature Measurement, Probe Number = 2, T = 28.000000 C 11:50:24 Temperature Measurement, Probe Number = 3, T = 28.255000 C 11:50:24 Temperature Measurement, Probe Number = 4, T = 28.655000 C 11:50:25 Temperature Measurement, Probe Number = 5, T = 24.507000 C 11:50:25 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.27 P3 = RTD Outside, Value = 29.02 P4 = Supply Water, Value = 28.00 P5 = Return Water, Value = 28.25 P6 = @ RTD Outside, Value = 28.66 P7 = Ambient, Value = 24.51 11:50:25 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000020, Bz = -0.000020 T 11:50:33 Move Stage, position = -0.274000 11:50:34 Stage Position Measurement, pos = -0.2744450 m 11:50:35 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:50:36 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:50:37 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:50:37 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:50:37 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 11:50:38 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:50:39 Main Volt Voltage Measurement, V = 0.000002 V 11:50:39 Temperature Measurement, Probe Number = 0, T = 29.265000 C 11:50:43 Temperature Measurement, Probe Number = 1, T = 29.022000 C 11:50:43 Temperature Measurement, Probe Number = 2, T = 27.998000 C 11:50:44 Temperature Measurement, Probe Number = 3, T = 28.249000 C 11:50:44 Temperature Measurement, Probe Number = 4, T = 28.653000 C 11:50:45 Temperature Measurement, Probe Number = 5, T = 24.456000 C 11:50:45 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.27 P3 = RTD Outside, Value = 29.02 P4 = Supply Water, Value = 28.00 P5 = Return Water, Value = 28.25 P6 = @ RTD Outside, Value = 28.65 P7 = Ambient, Value = 24.46 11:50:45 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000020, Bz = -0.000020 T 11:50:53 Move Stage, position = -0.272000 11:50:54 Stage Position Measurement, pos = -0.2724500 m 11:50:55 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:50:56 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:50:57 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:50:57 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:50:57 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 11:50:58 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:50:59 Main Volt Voltage Measurement, V = 0.000002 V 11:50:59 Temperature Measurement, Probe Number = 0, T = 29.261000 C 11:51:03 Temperature Measurement, Probe Number = 1, T = 29.023000 C 11:51:03 Temperature Measurement, Probe Number = 2, T = 27.982000 C 11:51:04 Temperature Measurement, Probe Number = 3, T = 28.248000 C 11:51:05 Temperature Measurement, Probe Number = 4, T = 28.652000 C 11:51:05 Temperature Measurement, Probe Number = 5, T = 24.510000 C 11:51:05 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.26 P3 = RTD Outside, Value = 29.02 P4 = Supply Water, Value = 27.98 P5 = Return Water, Value = 28.25 P6 = @ RTD Outside, Value = 28.65 P7 = Ambient, Value = 24.51 11:51:05 3 Axis Hall Probe Field Measurement, Bx = 0.000010, By = 0.000020, Bz = -0.000030 T 11:51:13 Move Stage, position = -0.270000 11:51:15 Stage Position Measurement, pos = -0.2704600 m 11:51:15 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:51:16 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:51:17 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:51:17 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:51:17 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 11:51:18 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:51:19 Main Volt Voltage Measurement, V = 0.000002 V 11:51:19 Temperature Measurement, Probe Number = 0, T = 29.264000 C 11:51:23 Temperature Measurement, Probe Number = 1, T = 29.017000 C 11:51:24 Temperature Measurement, Probe Number = 2, T = 27.993000 C 11:51:24 Temperature Measurement, Probe Number = 3, T = 28.256000 C 11:51:25 Temperature Measurement, Probe Number = 4, T = 28.650000 C 11:51:25 Temperature Measurement, Probe Number = 5, T = 24.548000 C 11:51:25 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.26 P3 = RTD Outside, Value = 29.02 P4 = Supply Water, Value = 27.99 P5 = Return Water, Value = 28.26 P6 = @ RTD Outside, Value = 28.65 P7 = Ambient, Value = 24.55 11:51:25 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000030, Bz = -0.000020 T 11:51:33 Move Stage, position = -0.268000 11:51:35 Stage Position Measurement, pos = -0.2684650 m 11:51:35 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:51:36 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:51:37 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:51:38 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:51:38 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 11:51:38 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:51:39 Main Volt Voltage Measurement, V = 0.000002 V 11:51:39 Temperature Measurement, Probe Number = 0, T = 29.266000 C 11:51:43 Temperature Measurement, Probe Number = 1, T = 29.019000 C 11:51:44 Temperature Measurement, Probe Number = 2, T = 28.028000 C 11:51:44 Temperature Measurement, Probe Number = 3, T = 28.261000 C 11:51:45 Temperature Measurement, Probe Number = 4, T = 28.647000 C 11:51:45 Temperature Measurement, Probe Number = 5, T = 24.559000 C 11:51:45 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.27 P3 = RTD Outside, Value = 29.02 P4 = Supply Water, Value = 28.03 P5 = Return Water, Value = 28.26 P6 = @ RTD Outside, Value = 28.65 P7 = Ambient, Value = 24.56 11:51:45 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000020, Bz = -0.000030 T 11:51:53 Move Stage, position = -0.266000 11:51:55 Stage Position Measurement, pos = -0.2664700 m 11:51:56 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:51:56 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:51:57 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:51:58 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:51:58 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 11:51:58 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:51:59 Main Volt Voltage Measurement, V = 0.000002 V 11:52:00 Temperature Measurement, Probe Number = 0, T = 29.270000 C 11:52:03 Temperature Measurement, Probe Number = 1, T = 29.021000 C 11:52:04 Temperature Measurement, Probe Number = 2, T = 28.057000 C 11:52:04 Temperature Measurement, Probe Number = 3, T = 28.281000 C 11:52:05 Temperature Measurement, Probe Number = 4, T = 28.646000 C 11:52:06 Temperature Measurement, Probe Number = 5, T = 24.714000 C 11:52:06 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.27 P3 = RTD Outside, Value = 29.02 P4 = Supply Water, Value = 28.06 P5 = Return Water, Value = 28.28 P6 = @ RTD Outside, Value = 28.65 P7 = Ambient, Value = 24.71 11:52:06 3 Axis Hall Probe Field Measurement, Bx = 0.000010, By = 0.000020, Bz = -0.000020 T 11:52:13 Move Stage, position = -0.264000 11:52:15 Stage Position Measurement, pos = -0.2644500 m 11:52:16 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:52:16 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:52:17 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:52:18 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:52:18 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 11:52:19 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:52:19 Main Volt Voltage Measurement, V = 0.000002 V 11:52:20 Temperature Measurement, Probe Number = 0, T = 29.270000 C 11:52:23 Temperature Measurement, Probe Number = 1, T = 29.007000 C 11:52:24 Temperature Measurement, Probe Number = 2, T = 28.034000 C 11:52:25 Temperature Measurement, Probe Number = 3, T = 28.312000 C 11:52:25 Temperature Measurement, Probe Number = 4, T = 28.626000 C 11:52:26 Temperature Measurement, Probe Number = 5, T = 24.704000 C 11:52:26 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.27 P3 = RTD Outside, Value = 29.01 P4 = Supply Water, Value = 28.03 P5 = Return Water, Value = 28.31 P6 = @ RTD Outside, Value = 28.63 P7 = Ambient, Value = 24.70 11:52:26 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000020, Bz = -0.000030 T 11:52:33 Move Stage, position = -0.262000 11:52:35 Stage Position Measurement, pos = -0.2624300 m 11:52:36 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:52:37 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:52:37 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:52:38 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:52:38 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 11:52:39 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:52:39 Main Volt Voltage Measurement, V = 0.000002 V 11:52:40 Temperature Measurement, Probe Number = 0, T = 29.277000 C 11:52:44 Temperature Measurement, Probe Number = 1, T = 29.012000 C 11:52:44 Temperature Measurement, Probe Number = 2, T = 28.029000 C 11:52:45 Temperature Measurement, Probe Number = 3, T = 28.346000 C 11:52:45 Temperature Measurement, Probe Number = 4, T = 28.606000 C 11:52:46 Temperature Measurement, Probe Number = 5, T = 24.725000 C 11:52:46 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.28 P3 = RTD Outside, Value = 29.01 P4 = Supply Water, Value = 28.03 P5 = Return Water, Value = 28.35 P6 = @ RTD Outside, Value = 28.61 P7 = Ambient, Value = 24.73 11:52:46 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000020, Bz = -0.000020 T 11:52:54 Move Stage, position = -0.260000 11:52:55 Stage Position Measurement, pos = -0.2604100 m 11:52:56 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:52:57 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:52:57 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:52:58 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:52:58 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 11:52:59 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:52:59 Main Volt Voltage Measurement, V = 0.000002 V 11:53:00 Temperature Measurement, Probe Number = 0, T = 29.287000 C 11:53:04 Temperature Measurement, Probe Number = 1, T = 29.029000 C 11:53:04 Temperature Measurement, Probe Number = 2, T = 28.042000 C 11:53:05 Temperature Measurement, Probe Number = 3, T = 28.373000 C 11:53:05 Temperature Measurement, Probe Number = 4, T = 28.612000 C 11:53:06 Temperature Measurement, Probe Number = 5, T = 24.771000 C 11:53:06 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.29 P3 = RTD Outside, Value = 29.03 P4 = Supply Water, Value = 28.04 P5 = Return Water, Value = 28.37 P6 = @ RTD Outside, Value = 28.61 P7 = Ambient, Value = 24.77 11:53:06 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000020, Bz = -0.000020 T 11:53:14 Move Stage, position = -0.258000 11:53:15 Stage Position Measurement, pos = -0.2584400 m 11:53:16 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:53:17 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:53:18 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:53:18 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:53:18 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 11:53:19 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:53:20 Main Volt Voltage Measurement, V = 0.000002 V 11:53:20 Temperature Measurement, Probe Number = 0, T = 29.292000 C 11:53:24 Temperature Measurement, Probe Number = 1, T = 29.037000 C 11:53:24 Temperature Measurement, Probe Number = 2, T = 28.057000 C 11:53:25 Temperature Measurement, Probe Number = 3, T = 28.385000 C 11:53:26 Temperature Measurement, Probe Number = 4, T = 28.622000 C 11:53:26 Temperature Measurement, Probe Number = 5, T = 24.797000 C 11:53:26 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.29 P3 = RTD Outside, Value = 29.04 P4 = Supply Water, Value = 28.06 P5 = Return Water, Value = 28.39 P6 = @ RTD Outside, Value = 28.62 P7 = Ambient, Value = 24.80 11:53:26 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000030, Bz = -0.000030 T 11:53:34 Move Stage, position = -0.256000 11:53:36 Stage Position Measurement, pos = -0.2564550 m 11:53:36 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:53:37 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:53:38 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:53:38 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:53:38 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 11:53:39 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:53:40 Main Volt Voltage Measurement, V = 0.000002 V 11:53:40 Temperature Measurement, Probe Number = 0, T = 29.293000 C 11:53:44 Temperature Measurement, Probe Number = 1, T = 29.045000 C 11:53:45 Temperature Measurement, Probe Number = 2, T = 28.091000 C 11:53:45 Temperature Measurement, Probe Number = 3, T = 28.411000 C 11:53:46 Temperature Measurement, Probe Number = 4, T = 28.633000 C 11:53:46 Temperature Measurement, Probe Number = 5, T = 24.787000 C 11:53:46 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.29 P3 = RTD Outside, Value = 29.05 P4 = Supply Water, Value = 28.09 P5 = Return Water, Value = 28.41 P6 = @ RTD Outside, Value = 28.63 P7 = Ambient, Value = 24.79 11:53:46 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000020, Bz = -0.000010 T 11:53:54 Move Stage, position = -0.254000 11:53:56 Stage Position Measurement, pos = -0.2544700 m 11:53:56 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:53:57 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:53:58 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:53:59 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:53:59 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 11:53:59 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:54:00 Main Volt Voltage Measurement, V = 0.000001 V 11:54:00 Temperature Measurement, Probe Number = 0, T = 29.297000 C 11:54:04 Temperature Measurement, Probe Number = 1, T = 29.053000 C 11:54:05 Temperature Measurement, Probe Number = 2, T = 28.130000 C 11:54:05 Temperature Measurement, Probe Number = 3, T = 28.446000 C 11:54:06 Temperature Measurement, Probe Number = 4, T = 28.645000 C 11:54:06 Temperature Measurement, Probe Number = 5, T = 24.808000 C 11:54:06 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.30 P3 = RTD Outside, Value = 29.05 P4 = Supply Water, Value = 28.13 P5 = Return Water, Value = 28.45 P6 = @ RTD Outside, Value = 28.64 P7 = Ambient, Value = 24.81 11:54:06 3 Axis Hall Probe Field Measurement, Bx = 0.000010, By = 0.000020, Bz = -0.000020 T 11:54:14 Move Stage, position = -0.252000 11:54:16 Stage Position Measurement, pos = -0.2524500 m 11:54:16 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:54:17 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:54:18 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:54:19 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:54:19 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 11:54:19 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:54:20 Main Volt Voltage Measurement, V = 0.000002 V 11:54:21 Temperature Measurement, Probe Number = 0, T = 29.301000 C 11:54:24 Temperature Measurement, Probe Number = 1, T = 29.060000 C 11:54:25 Temperature Measurement, Probe Number = 2, T = 28.109000 C 11:54:25 Temperature Measurement, Probe Number = 3, T = 28.457000 C 11:54:26 Temperature Measurement, Probe Number = 4, T = 28.658000 C 11:54:26 Temperature Measurement, Probe Number = 5, T = 24.810000 C 11:54:26 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.30 P3 = RTD Outside, Value = 29.06 P4 = Supply Water, Value = 28.11 P5 = Return Water, Value = 28.46 P6 = @ RTD Outside, Value = 28.66 P7 = Ambient, Value = 24.81 11:54:27 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000030, Bz = -0.000030 T 11:54:34 Move Stage, position = -0.250000 11:54:36 Stage Position Measurement, pos = -0.2504500 m 11:54:37 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:54:37 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:54:38 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:54:39 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:54:39 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 11:54:39 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:54:40 Main Volt Voltage Measurement, V = 0.000002 V 11:54:41 Temperature Measurement, Probe Number = 0, T = 29.305000 C 11:54:44 Temperature Measurement, Probe Number = 1, T = 29.063000 C 11:54:45 Temperature Measurement, Probe Number = 2, T = 28.087000 C 11:54:45 Temperature Measurement, Probe Number = 3, T = 28.468000 C 11:54:46 Temperature Measurement, Probe Number = 4, T = 28.668000 C 11:54:47 Temperature Measurement, Probe Number = 5, T = 24.805000 C 11:54:47 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.30 P3 = RTD Outside, Value = 29.06 P4 = Supply Water, Value = 28.09 P5 = Return Water, Value = 28.47 P6 = @ RTD Outside, Value = 28.67 P7 = Ambient, Value = 24.80 11:54:47 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000030, Bz = -0.000020 T 11:54:54 Move Stage, position = -0.248000 11:54:56 Stage Position Measurement, pos = -0.2484350 m 11:54:57 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:54:57 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:54:58 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:54:59 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:54:59 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 11:55:00 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:55:00 Main Volt Voltage Measurement, V = 0.000001 V 11:55:01 Temperature Measurement, Probe Number = 0, T = 29.307000 C 11:55:04 Temperature Measurement, Probe Number = 1, T = 29.060000 C 11:55:05 Temperature Measurement, Probe Number = 2, T = 28.112000 C 11:55:06 Temperature Measurement, Probe Number = 3, T = 28.490000 C 11:55:06 Temperature Measurement, Probe Number = 4, T = 28.672000 C 11:55:07 Temperature Measurement, Probe Number = 5, T = 24.797000 C 11:55:07 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.31 P3 = RTD Outside, Value = 29.06 P4 = Supply Water, Value = 28.11 P5 = Return Water, Value = 28.49 P6 = @ RTD Outside, Value = 28.67 P7 = Ambient, Value = 24.80 11:55:07 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000020, Bz = -0.000020 T 11:55:14 Move Stage, position = -0.246000 11:55:16 Stage Position Measurement, pos = -0.2464350 m 11:55:17 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:55:18 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:55:18 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:55:19 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:55:19 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 11:55:20 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:55:20 Main Volt Voltage Measurement, V = 0.000001 V 11:55:21 Temperature Measurement, Probe Number = 0, T = 29.309000 C 11:55:24 Temperature Measurement, Probe Number = 1, T = 29.061000 C 11:55:25 Temperature Measurement, Probe Number = 2, T = 28.165000 C 11:55:26 Temperature Measurement, Probe Number = 3, T = 28.510000 C 11:55:26 Temperature Measurement, Probe Number = 4, T = 28.672000 C 11:55:27 Temperature Measurement, Probe Number = 5, T = 24.789000 C 11:55:27 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.31 P3 = RTD Outside, Value = 29.06 P4 = Supply Water, Value = 28.16 P5 = Return Water, Value = 28.51 P6 = @ RTD Outside, Value = 28.67 P7 = Ambient, Value = 24.79 11:55:27 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000040, Bz = -0.000020 T 11:55:35 Move Stage, position = -0.244000 11:55:36 Stage Position Measurement, pos = -0.2444200 m 11:55:37 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:55:38 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:55:38 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:55:39 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:55:39 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 11:55:40 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:55:40 Main Volt Voltage Measurement, V = 0.000002 V 11:55:41 Temperature Measurement, Probe Number = 0, T = 29.312000 C 11:55:45 Temperature Measurement, Probe Number = 1, T = 29.068000 C 11:55:45 Temperature Measurement, Probe Number = 2, T = 28.195000 C 11:55:46 Temperature Measurement, Probe Number = 3, T = 28.534000 C 11:55:46 Temperature Measurement, Probe Number = 4, T = 28.679000 C 11:55:47 Temperature Measurement, Probe Number = 5, T = 24.800000 C 11:55:47 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.31 P3 = RTD Outside, Value = 29.07 P4 = Supply Water, Value = 28.20 P5 = Return Water, Value = 28.53 P6 = @ RTD Outside, Value = 28.68 P7 = Ambient, Value = 24.80 11:55:47 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000030, Bz = -0.000010 T 11:55:55 Move Stage, position = -0.242000 11:55:56 Stage Position Measurement, pos = -0.2423800 m 11:55:57 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:55:58 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:55:58 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:55:59 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:55:59 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 11:56:00 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:56:00 Main Volt Voltage Measurement, V = 0.000002 V 11:56:01 Temperature Measurement, Probe Number = 0, T = 29.314000 C 11:56:05 Temperature Measurement, Probe Number = 1, T = 29.069000 C 11:56:05 Temperature Measurement, Probe Number = 2, T = 28.182000 C 11:56:06 Temperature Measurement, Probe Number = 3, T = 28.509000 C 11:56:06 Temperature Measurement, Probe Number = 4, T = 28.690000 C 11:56:07 Temperature Measurement, Probe Number = 5, T = 24.738000 C 11:56:07 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.31 P3 = RTD Outside, Value = 29.07 P4 = Supply Water, Value = 28.18 P5 = Return Water, Value = 28.51 P6 = @ RTD Outside, Value = 28.69 P7 = Ambient, Value = 24.74 11:56:07 3 Axis Hall Probe Field Measurement, Bx = -0.000010, By = 0.000020, Bz = -0.000020 T 11:56:15 Move Stage, position = -0.240000 11:56:16 Stage Position Measurement, pos = -0.2404100 m 11:56:17 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:56:18 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:56:19 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:56:19 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:56:19 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 11:56:20 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:56:21 Main Volt Voltage Measurement, V = 0.000002 V 11:56:21 Temperature Measurement, Probe Number = 0, T = 29.318000 C 11:56:25 Temperature Measurement, Probe Number = 1, T = 29.075000 C 11:56:25 Temperature Measurement, Probe Number = 2, T = 28.174000 C 11:56:26 Temperature Measurement, Probe Number = 3, T = 28.482000 C 11:56:27 Temperature Measurement, Probe Number = 4, T = 28.698000 C 11:56:27 Temperature Measurement, Probe Number = 5, T = 24.740000 C 11:56:27 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.32 P3 = RTD Outside, Value = 29.07 P4 = Supply Water, Value = 28.17 P5 = Return Water, Value = 28.48 P6 = @ RTD Outside, Value = 28.70 P7 = Ambient, Value = 24.74 11:56:27 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000020, Bz = -0.000020 T 11:56:35 Move Stage, position = -0.238000 11:56:36 Stage Position Measurement, pos = -0.2384200 m 11:56:37 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:56:38 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:56:39 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:56:39 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:56:39 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 11:56:40 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:56:41 Main Volt Voltage Measurement, V = 0.000002 V 11:56:41 Temperature Measurement, Probe Number = 0, T = 29.314000 C 11:56:45 Temperature Measurement, Probe Number = 1, T = 29.069000 C 11:56:45 Temperature Measurement, Probe Number = 2, T = 28.174000 C 11:56:46 Temperature Measurement, Probe Number = 3, T = 28.460000 C 11:56:47 Temperature Measurement, Probe Number = 4, T = 28.699000 C 11:56:47 Temperature Measurement, Probe Number = 5, T = 24.700000 C 11:56:47 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.31 P3 = RTD Outside, Value = 29.07 P4 = Supply Water, Value = 28.17 P5 = Return Water, Value = 28.46 P6 = @ RTD Outside, Value = 28.70 P7 = Ambient, Value = 24.70 11:56:47 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000020, Bz = -0.000030 T 11:56:55 Move Stage, position = -0.236000 11:56:57 Stage Position Measurement, pos = -0.2364450 m 11:56:57 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:56:58 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:56:59 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:57:00 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:57:00 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 11:57:00 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:57:01 Main Volt Voltage Measurement, V = 0.000002 V 11:57:01 Temperature Measurement, Probe Number = 0, T = 29.310000 C 11:57:05 Temperature Measurement, Probe Number = 1, T = 29.068000 C 11:57:06 Temperature Measurement, Probe Number = 2, T = 28.186000 C 11:57:06 Temperature Measurement, Probe Number = 3, T = 28.466000 C 11:57:07 Temperature Measurement, Probe Number = 4, T = 28.699000 C 11:57:07 Temperature Measurement, Probe Number = 5, T = 24.797000 C 11:57:07 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.31 P3 = RTD Outside, Value = 29.07 P4 = Supply Water, Value = 28.19 P5 = Return Water, Value = 28.47 P6 = @ RTD Outside, Value = 28.70 P7 = Ambient, Value = 24.80 11:57:07 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000020, Bz = -0.000020 T 11:57:15 Move Stage, position = -0.234000 11:57:17 Stage Position Measurement, pos = -0.2344350 m 11:57:17 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:57:18 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:57:19 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:57:20 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:57:20 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 11:57:20 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:57:21 Main Volt Voltage Measurement, V = 0.000002 V 11:57:21 Temperature Measurement, Probe Number = 0, T = 29.310000 C 11:57:25 Temperature Measurement, Probe Number = 1, T = 29.069000 C 11:57:26 Temperature Measurement, Probe Number = 2, T = 28.212000 C 11:57:26 Temperature Measurement, Probe Number = 3, T = 28.442000 C 11:57:27 Temperature Measurement, Probe Number = 4, T = 28.698000 C 11:57:27 Temperature Measurement, Probe Number = 5, T = 24.773000 C 11:57:27 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.31 P3 = RTD Outside, Value = 29.07 P4 = Supply Water, Value = 28.21 P5 = Return Water, Value = 28.44 P6 = @ RTD Outside, Value = 28.70 P7 = Ambient, Value = 24.77 11:57:28 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000030, Bz = -0.000020 T 11:57:35 Move Stage, position = -0.232000 11:57:37 Stage Position Measurement, pos = -0.2324300 m 11:57:38 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:57:38 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:57:39 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:57:40 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:57:40 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 11:57:40 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:57:41 Main Volt Voltage Measurement, V = 0.000002 V 11:57:42 Temperature Measurement, Probe Number = 0, T = 29.311000 C 11:57:45 Temperature Measurement, Probe Number = 1, T = 29.071000 C 11:57:46 Temperature Measurement, Probe Number = 2, T = 28.220000 C 11:57:46 Temperature Measurement, Probe Number = 3, T = 28.395000 C 11:57:47 Temperature Measurement, Probe Number = 4, T = 28.698000 C 11:57:48 Temperature Measurement, Probe Number = 5, T = 24.742000 C 11:57:48 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.31 P3 = RTD Outside, Value = 29.07 P4 = Supply Water, Value = 28.22 P5 = Return Water, Value = 28.39 P6 = @ RTD Outside, Value = 28.70 P7 = Ambient, Value = 24.74 11:57:48 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000020, Bz = -0.000020 T 11:57:55 Move Stage, position = -0.230000 11:57:57 Stage Position Measurement, pos = -0.2304100 m 11:57:58 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:57:58 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:57:59 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:58:00 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:58:00 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 11:58:01 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:58:01 Main Volt Voltage Measurement, V = 0.000002 V 11:58:02 Temperature Measurement, Probe Number = 0, T = 29.316000 C 11:58:06 Temperature Measurement, Probe Number = 1, T = 29.073000 C 11:58:06 Temperature Measurement, Probe Number = 2, T = 28.203000 C 11:58:07 Temperature Measurement, Probe Number = 3, T = 28.407000 C 11:58:07 Temperature Measurement, Probe Number = 4, T = 28.700000 C 11:58:08 Temperature Measurement, Probe Number = 5, T = 24.781000 C 11:58:08 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.32 P3 = RTD Outside, Value = 29.07 P4 = Supply Water, Value = 28.20 P5 = Return Water, Value = 28.41 P6 = @ RTD Outside, Value = 28.70 P7 = Ambient, Value = 24.78 11:58:08 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000030, Bz = -0.000030 T 11:58:16 Move Stage, position = -0.228000 11:58:17 Stage Position Measurement, pos = -0.2284050 m 11:58:18 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:58:19 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:58:19 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:58:20 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:58:20 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 11:58:21 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:58:21 Main Volt Voltage Measurement, V = 0.000002 V 11:58:22 Temperature Measurement, Probe Number = 0, T = 29.317000 C 11:58:26 Temperature Measurement, Probe Number = 1, T = 29.075000 C 11:58:26 Temperature Measurement, Probe Number = 2, T = 28.156000 C 11:58:27 Temperature Measurement, Probe Number = 3, T = 28.427000 C 11:58:27 Temperature Measurement, Probe Number = 4, T = 28.703000 C 11:58:28 Temperature Measurement, Probe Number = 5, T = 24.803000 C 11:58:28 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.32 P3 = RTD Outside, Value = 29.07 P4 = Supply Water, Value = 28.16 P5 = Return Water, Value = 28.43 P6 = @ RTD Outside, Value = 28.70 P7 = Ambient, Value = 24.80 11:58:28 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000030, Bz = -0.000020 T 11:58:36 Move Stage, position = -0.226000 11:58:37 Stage Position Measurement, pos = -0.2264050 m 11:58:38 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:58:39 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:58:40 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:58:40 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:58:40 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 11:58:41 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:58:42 Main Volt Voltage Measurement, V = 0.000002 V 11:58:42 Temperature Measurement, Probe Number = 0, T = 29.318000 C 11:58:46 Temperature Measurement, Probe Number = 1, T = 29.077000 C 11:58:46 Temperature Measurement, Probe Number = 2, T = 28.088000 C 11:58:47 Temperature Measurement, Probe Number = 3, T = 28.422000 C 11:58:48 Temperature Measurement, Probe Number = 4, T = 28.702000 C 11:58:48 Temperature Measurement, Probe Number = 5, T = 24.766000 C 11:58:48 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.32 P3 = RTD Outside, Value = 29.08 P4 = Supply Water, Value = 28.09 P5 = Return Water, Value = 28.42 P6 = @ RTD Outside, Value = 28.70 P7 = Ambient, Value = 24.77 11:58:48 3 Axis Hall Probe Field Measurement, Bx = 0.000010, By = 0.000030, Bz = -0.000020 T 11:58:56 Move Stage, position = -0.224000 11:58:58 Stage Position Measurement, pos = -0.2243700 m 11:58:58 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:58:59 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:59:00 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:59:00 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:59:00 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 11:59:01 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:59:02 Main Volt Voltage Measurement, V = 0.000001 V 11:59:02 Temperature Measurement, Probe Number = 0, T = 29.317000 C 11:59:06 Temperature Measurement, Probe Number = 1, T = 29.071000 C 11:59:06 Temperature Measurement, Probe Number = 2, T = 28.022000 C 11:59:07 Temperature Measurement, Probe Number = 3, T = 28.418000 C 11:59:08 Temperature Measurement, Probe Number = 4, T = 28.699000 C 11:59:08 Temperature Measurement, Probe Number = 5, T = 24.691000 C 11:59:08 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.32 P3 = RTD Outside, Value = 29.07 P4 = Supply Water, Value = 28.02 P5 = Return Water, Value = 28.42 P6 = @ RTD Outside, Value = 28.70 P7 = Ambient, Value = 24.69 11:59:08 3 Axis Hall Probe Field Measurement, Bx = 0.000010, By = 0.000020, Bz = -0.000020 T 11:59:16 Move Stage, position = -0.222000 11:59:18 Stage Position Measurement, pos = -0.2223800 m 11:59:18 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:59:19 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:59:20 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:59:20 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:59:20 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 11:59:21 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:59:22 Main Volt Voltage Measurement, V = 0.000002 V 11:59:22 Temperature Measurement, Probe Number = 0, T = 29.316000 C 11:59:26 Temperature Measurement, Probe Number = 1, T = 29.073000 C 11:59:27 Temperature Measurement, Probe Number = 2, T = 28.015000 C 11:59:27 Temperature Measurement, Probe Number = 3, T = 28.435000 C 11:59:28 Temperature Measurement, Probe Number = 4, T = 28.695000 C 11:59:28 Temperature Measurement, Probe Number = 5, T = 24.727000 C 11:59:28 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.32 P3 = RTD Outside, Value = 29.07 P4 = Supply Water, Value = 28.02 P5 = Return Water, Value = 28.43 P6 = @ RTD Outside, Value = 28.70 P7 = Ambient, Value = 24.73 11:59:28 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000030, Bz = -0.000020 T 11:59:36 Move Stage, position = -0.220000 11:59:38 Stage Position Measurement, pos = -0.2203850 m 11:59:38 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:59:39 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:59:40 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:59:41 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:59:41 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 11:59:41 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:59:42 Main Volt Voltage Measurement, V = 0.000002 V 11:59:42 Temperature Measurement, Probe Number = 0, T = 29.320000 C 11:59:46 Temperature Measurement, Probe Number = 1, T = 29.078000 C 11:59:47 Temperature Measurement, Probe Number = 2, T = 28.070000 C 11:59:47 Temperature Measurement, Probe Number = 3, T = 28.447000 C 11:59:48 Temperature Measurement, Probe Number = 4, T = 28.698000 C 11:59:48 Temperature Measurement, Probe Number = 5, T = 24.758000 C 11:59:48 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.32 P3 = RTD Outside, Value = 29.08 P4 = Supply Water, Value = 28.07 P5 = Return Water, Value = 28.45 P6 = @ RTD Outside, Value = 28.70 P7 = Ambient, Value = 24.76 11:59:48 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000020, Bz = -0.000020 T 11:59:56 Move Stage, position = -0.218000 11:59:58 Stage Position Measurement, pos = -0.2184100 m 11:59:59 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 11:59:59 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:00:00 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:00:01 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:00:01 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:00:01 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:00:02 Main Volt Voltage Measurement, V = 0.000002 V 12:00:03 Temperature Measurement, Probe Number = 0, T = 29.323000 C 12:00:06 Temperature Measurement, Probe Number = 1, T = 29.082000 C 12:00:07 Temperature Measurement, Probe Number = 2, T = 28.131000 C 12:00:07 Temperature Measurement, Probe Number = 3, T = 28.455000 C 12:00:08 Temperature Measurement, Probe Number = 4, T = 28.701000 C 12:00:09 Temperature Measurement, Probe Number = 5, T = 24.803000 C 12:00:09 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.32 P3 = RTD Outside, Value = 29.08 P4 = Supply Water, Value = 28.13 P5 = Return Water, Value = 28.45 P6 = @ RTD Outside, Value = 28.70 P7 = Ambient, Value = 24.80 12:00:09 3 Axis Hall Probe Field Measurement, Bx = 0.000010, By = 0.000030, Bz = -0.000030 T 12:00:16 Move Stage, position = -0.216000 12:00:18 Stage Position Measurement, pos = -0.2164100 m 12:00:19 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:00:19 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:00:20 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:00:21 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:00:21 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:00:21 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:00:22 Main Volt Voltage Measurement, V = 0.000002 V 12:00:23 Temperature Measurement, Probe Number = 0, T = 29.328000 C 12:00:26 Temperature Measurement, Probe Number = 1, T = 29.083000 C 12:00:27 Temperature Measurement, Probe Number = 2, T = 28.191000 C 12:00:27 Temperature Measurement, Probe Number = 3, T = 28.482000 C 12:00:28 Temperature Measurement, Probe Number = 4, T = 28.705000 C 12:00:29 Temperature Measurement, Probe Number = 5, T = 24.890000 C 12:00:29 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.33 P3 = RTD Outside, Value = 29.08 P4 = Supply Water, Value = 28.19 P5 = Return Water, Value = 28.48 P6 = @ RTD Outside, Value = 28.70 P7 = Ambient, Value = 24.89 12:00:29 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000030, Bz = -0.000030 T 12:00:36 Move Stage, position = -0.214000 12:00:38 Stage Position Measurement, pos = -0.2144050 m 12:00:39 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:00:39 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:00:40 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:00:41 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:00:41 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:00:42 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:00:42 Main Volt Voltage Measurement, V = 0.000002 V 12:00:43 Temperature Measurement, Probe Number = 0, T = 29.331000 C 12:00:46 Temperature Measurement, Probe Number = 1, T = 29.086000 C 12:00:47 Temperature Measurement, Probe Number = 2, T = 28.244000 C 12:00:48 Temperature Measurement, Probe Number = 3, T = 28.513000 C 12:00:48 Temperature Measurement, Probe Number = 4, T = 28.708000 C 12:00:49 Temperature Measurement, Probe Number = 5, T = 24.923000 C 12:00:49 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.33 P3 = RTD Outside, Value = 29.09 P4 = Supply Water, Value = 28.24 P5 = Return Water, Value = 28.51 P6 = @ RTD Outside, Value = 28.71 P7 = Ambient, Value = 24.92 12:00:49 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000030, Bz = -0.000020 T 12:00:57 Move Stage, position = -0.212000 12:00:58 Stage Position Measurement, pos = -0.2123750 m 12:00:59 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:01:00 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:01:00 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:01:01 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:01:01 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:01:02 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:01:02 Main Volt Voltage Measurement, V = 0.000002 V 12:01:03 Temperature Measurement, Probe Number = 0, T = 29.337000 C 12:01:07 Temperature Measurement, Probe Number = 1, T = 29.092000 C 12:01:07 Temperature Measurement, Probe Number = 2, T = 28.277000 C 12:01:08 Temperature Measurement, Probe Number = 3, T = 28.537000 C 12:01:08 Temperature Measurement, Probe Number = 4, T = 28.712000 C 12:01:09 Temperature Measurement, Probe Number = 5, T = 24.957000 C 12:01:09 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.34 P3 = RTD Outside, Value = 29.09 P4 = Supply Water, Value = 28.28 P5 = Return Water, Value = 28.54 P6 = @ RTD Outside, Value = 28.71 P7 = Ambient, Value = 24.96 12:01:09 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000020, Bz = -0.000020 T 12:01:17 Move Stage, position = -0.210000 12:01:18 Stage Position Measurement, pos = -0.2103650 m 12:01:19 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:01:20 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:01:20 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:01:21 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:01:21 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:01:22 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:01:22 Main Volt Voltage Measurement, V = 0.000001 V 12:01:23 Temperature Measurement, Probe Number = 0, T = 29.339000 C 12:01:27 Temperature Measurement, Probe Number = 1, T = 29.095000 C 12:01:27 Temperature Measurement, Probe Number = 2, T = 28.258000 C 12:01:28 Temperature Measurement, Probe Number = 3, T = 28.549000 C 12:01:28 Temperature Measurement, Probe Number = 4, T = 28.719000 C 12:01:29 Temperature Measurement, Probe Number = 5, T = 24.967000 C 12:01:29 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.34 P3 = RTD Outside, Value = 29.09 P4 = Supply Water, Value = 28.26 P5 = Return Water, Value = 28.55 P6 = @ RTD Outside, Value = 28.72 P7 = Ambient, Value = 24.97 12:01:29 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000020, Bz = -0.000030 T 12:01:37 Move Stage, position = -0.208000 12:01:38 Stage Position Measurement, pos = -0.2083500 m 12:01:39 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:01:40 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:01:41 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:01:41 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:01:41 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:01:42 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:01:42 Main Volt Voltage Measurement, V = 0.000001 V 12:01:43 Temperature Measurement, Probe Number = 0, T = 29.341000 C 12:01:47 Temperature Measurement, Probe Number = 1, T = 29.103000 C 12:01:47 Temperature Measurement, Probe Number = 2, T = 28.213000 C 12:01:48 Temperature Measurement, Probe Number = 3, T = 28.535000 C 12:01:48 Temperature Measurement, Probe Number = 4, T = 28.726000 C 12:01:49 Temperature Measurement, Probe Number = 5, T = 24.935000 C 12:01:49 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.34 P3 = RTD Outside, Value = 29.10 P4 = Supply Water, Value = 28.21 P5 = Return Water, Value = 28.54 P6 = @ RTD Outside, Value = 28.73 P7 = Ambient, Value = 24.93 12:01:49 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000020, Bz = -0.000020 T 12:01:57 Move Stage, position = -0.206000 12:01:58 Stage Position Measurement, pos = -0.2063450 m 12:01:59 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:02:00 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:02:01 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:02:01 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:02:01 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:02:02 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:02:03 Main Volt Voltage Measurement, V = 0.000001 V 12:02:03 Temperature Measurement, Probe Number = 0, T = 29.342000 C 12:02:07 Temperature Measurement, Probe Number = 1, T = 29.101000 C 12:02:07 Temperature Measurement, Probe Number = 2, T = 28.197000 C 12:02:08 Temperature Measurement, Probe Number = 3, T = 28.543000 C 12:02:09 Temperature Measurement, Probe Number = 4, T = 28.730000 C 12:02:09 Temperature Measurement, Probe Number = 5, T = 24.953000 C 12:02:09 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.34 P3 = RTD Outside, Value = 29.10 P4 = Supply Water, Value = 28.20 P5 = Return Water, Value = 28.54 P6 = @ RTD Outside, Value = 28.73 P7 = Ambient, Value = 24.95 12:02:09 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000020, Bz = -0.000030 T 12:02:17 Move Stage, position = -0.204000 12:02:19 Stage Position Measurement, pos = -0.2043650 m 12:02:19 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:02:20 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:02:21 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:02:21 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:02:21 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:02:22 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:02:23 Main Volt Voltage Measurement, V = 0.000001 V 12:02:23 Temperature Measurement, Probe Number = 0, T = 29.341000 C 12:02:27 Temperature Measurement, Probe Number = 1, T = 29.102000 C 12:02:28 Temperature Measurement, Probe Number = 2, T = 28.179000 C 12:02:28 Temperature Measurement, Probe Number = 3, T = 28.551000 C 12:02:29 Temperature Measurement, Probe Number = 4, T = 28.733000 C 12:02:29 Temperature Measurement, Probe Number = 5, T = 24.951000 C 12:02:29 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.34 P3 = RTD Outside, Value = 29.10 P4 = Supply Water, Value = 28.18 P5 = Return Water, Value = 28.55 P6 = @ RTD Outside, Value = 28.73 P7 = Ambient, Value = 24.95 12:02:29 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000030, Bz = -0.000020 T 12:02:37 Move Stage, position = -0.202000 12:02:39 Stage Position Measurement, pos = -0.2023450 m 12:02:39 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:02:40 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:02:41 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:02:42 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:02:42 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:02:42 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:02:43 Main Volt Voltage Measurement, V = 0.000001 V 12:02:43 Temperature Measurement, Probe Number = 0, T = 29.343000 C 12:02:47 Temperature Measurement, Probe Number = 1, T = 29.103000 C 12:02:48 Temperature Measurement, Probe Number = 2, T = 28.145000 C 12:02:48 Temperature Measurement, Probe Number = 3, T = 28.549000 C 12:02:49 Temperature Measurement, Probe Number = 4, T = 28.736000 C 12:02:49 Temperature Measurement, Probe Number = 5, T = 24.925000 C 12:02:49 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.34 P3 = RTD Outside, Value = 29.10 P4 = Supply Water, Value = 28.14 P5 = Return Water, Value = 28.55 P6 = @ RTD Outside, Value = 28.74 P7 = Ambient, Value = 24.93 12:02:49 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000020, Bz = -0.000020 T 12:02:57 Move Stage, position = -0.200000 12:02:59 Stage Position Measurement, pos = -0.2003700 m 12:02:59 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:03:00 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:03:01 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:03:02 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:03:02 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:03:02 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:03:03 Main Volt Voltage Measurement, V = 0.000001 V 12:03:04 Temperature Measurement, Probe Number = 0, T = 29.344000 C 12:03:07 Temperature Measurement, Probe Number = 1, T = 29.105000 C 12:03:08 Temperature Measurement, Probe Number = 2, T = 28.091000 C 12:03:08 Temperature Measurement, Probe Number = 3, T = 28.528000 C 12:03:09 Temperature Measurement, Probe Number = 4, T = 28.737000 C 12:03:09 Temperature Measurement, Probe Number = 5, T = 24.913000 C 12:03:09 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.34 P3 = RTD Outside, Value = 29.11 P4 = Supply Water, Value = 28.09 P5 = Return Water, Value = 28.53 P6 = @ RTD Outside, Value = 28.74 P7 = Ambient, Value = 24.91 12:03:10 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000030, Bz = -0.000020 T 12:03:17 Move Stage, position = -0.198000 12:03:19 Stage Position Measurement, pos = -0.1983650 m 12:03:20 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:03:20 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:03:21 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:03:22 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:03:22 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:03:22 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:03:23 Main Volt Voltage Measurement, V = 0.000001 V 12:03:24 Temperature Measurement, Probe Number = 0, T = 29.347000 C 12:03:27 Temperature Measurement, Probe Number = 1, T = 29.107000 C 12:03:28 Temperature Measurement, Probe Number = 2, T = 28.053000 C 12:03:28 Temperature Measurement, Probe Number = 3, T = 28.517000 C 12:03:29 Temperature Measurement, Probe Number = 4, T = 28.740000 C 12:03:30 Temperature Measurement, Probe Number = 5, T = 24.956000 C 12:03:30 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.35 P3 = RTD Outside, Value = 29.11 P4 = Supply Water, Value = 28.05 P5 = Return Water, Value = 28.52 P6 = @ RTD Outside, Value = 28.74 P7 = Ambient, Value = 24.96 12:03:30 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000030, Bz = -0.000020 T 12:03:37 Move Stage, position = -0.196000 12:03:39 Stage Position Measurement, pos = -0.1963700 m 12:03:40 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:03:40 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:03:41 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:03:42 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:03:42 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:03:43 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:03:43 Main Volt Voltage Measurement, V = 0.000001 V 12:03:44 Temperature Measurement, Probe Number = 0, T = 29.346000 C 12:03:47 Temperature Measurement, Probe Number = 1, T = 29.112000 C 12:03:48 Temperature Measurement, Probe Number = 2, T = 28.039000 C 12:03:49 Temperature Measurement, Probe Number = 3, T = 28.508000 C 12:03:49 Temperature Measurement, Probe Number = 4, T = 28.744000 C 12:03:50 Temperature Measurement, Probe Number = 5, T = 24.989000 C 12:03:50 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.35 P3 = RTD Outside, Value = 29.11 P4 = Supply Water, Value = 28.04 P5 = Return Water, Value = 28.51 P6 = @ RTD Outside, Value = 28.74 P7 = Ambient, Value = 24.99 12:03:50 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000030, Bz = -0.000010 T 12:03:57 Move Stage, position = -0.194000 12:03:59 Stage Position Measurement, pos = -0.1943450 m 12:04:00 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:04:01 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:04:01 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:04:02 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:04:02 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:04:03 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:04:03 Main Volt Voltage Measurement, V = 0.000001 V 12:04:04 Temperature Measurement, Probe Number = 0, T = 29.348000 C 12:04:07 Temperature Measurement, Probe Number = 1, T = 29.114000 C 12:04:08 Temperature Measurement, Probe Number = 2, T = 28.004000 C 12:04:09 Temperature Measurement, Probe Number = 3, T = 28.488000 C 12:04:09 Temperature Measurement, Probe Number = 4, T = 28.747000 C 12:04:10 Temperature Measurement, Probe Number = 5, T = 24.943000 C 12:04:10 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.35 P3 = RTD Outside, Value = 29.11 P4 = Supply Water, Value = 28.00 P5 = Return Water, Value = 28.49 P6 = @ RTD Outside, Value = 28.75 P7 = Ambient, Value = 24.94 12:04:10 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000020, Bz = -0.000010 T 12:04:18 Move Stage, position = -0.192000 12:04:19 Stage Position Measurement, pos = -0.1923200 m 12:04:20 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:04:21 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:04:21 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:04:22 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:04:22 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:04:23 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:04:23 Main Volt Voltage Measurement, V = 0.000001 V 12:04:24 Temperature Measurement, Probe Number = 0, T = 29.350000 C 12:04:28 Temperature Measurement, Probe Number = 1, T = 29.113000 C 12:04:28 Temperature Measurement, Probe Number = 2, T = 27.978000 C 12:04:29 Temperature Measurement, Probe Number = 3, T = 28.484000 C 12:04:29 Temperature Measurement, Probe Number = 4, T = 28.748000 C 12:04:30 Temperature Measurement, Probe Number = 5, T = 24.947000 C 12:04:30 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.35 P3 = RTD Outside, Value = 29.11 P4 = Supply Water, Value = 27.98 P5 = Return Water, Value = 28.48 P6 = @ RTD Outside, Value = 28.75 P7 = Ambient, Value = 24.95 12:04:30 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000020, Bz = -0.000030 T 12:04:38 Move Stage, position = -0.190000 12:04:39 Stage Position Measurement, pos = -0.1903150 m 12:04:40 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:04:41 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:04:41 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:04:42 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:04:42 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:04:43 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:04:43 Main Volt Voltage Measurement, V = 0.000001 V 12:04:44 Temperature Measurement, Probe Number = 0, T = 29.349000 C 12:04:48 Temperature Measurement, Probe Number = 1, T = 29.112000 C 12:04:48 Temperature Measurement, Probe Number = 2, T = 27.960000 C 12:04:49 Temperature Measurement, Probe Number = 3, T = 28.490000 C 12:04:49 Temperature Measurement, Probe Number = 4, T = 28.748000 C 12:04:50 Temperature Measurement, Probe Number = 5, T = 24.915000 C 12:04:50 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.35 P3 = RTD Outside, Value = 29.11 P4 = Supply Water, Value = 27.96 P5 = Return Water, Value = 28.49 P6 = @ RTD Outside, Value = 28.75 P7 = Ambient, Value = 24.91 12:04:50 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000020, Bz = -0.000030 T 12:04:58 Move Stage, position = -0.188000 12:04:59 Stage Position Measurement, pos = -0.1883000 m 12:05:00 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:05:01 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:05:02 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:05:02 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:05:02 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:05:03 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:05:04 Main Volt Voltage Measurement, V = 0.000001 V 12:05:04 Temperature Measurement, Probe Number = 0, T = 29.350000 C 12:05:08 Temperature Measurement, Probe Number = 1, T = 29.114000 C 12:05:08 Temperature Measurement, Probe Number = 2, T = 27.964000 C 12:05:09 Temperature Measurement, Probe Number = 3, T = 28.499000 C 12:05:10 Temperature Measurement, Probe Number = 4, T = 28.749000 C 12:05:10 Temperature Measurement, Probe Number = 5, T = 24.872000 C 12:05:10 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.35 P3 = RTD Outside, Value = 29.11 P4 = Supply Water, Value = 27.96 P5 = Return Water, Value = 28.50 P6 = @ RTD Outside, Value = 28.75 P7 = Ambient, Value = 24.87 12:05:10 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000020, Bz = -0.000030 T 12:05:18 Move Stage, position = -0.186000 12:05:20 Stage Position Measurement, pos = -0.1863200 m 12:05:20 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:05:21 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:05:22 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:05:22 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:05:23 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:05:23 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:05:24 Main Volt Voltage Measurement, V = 0.000001 V 12:05:24 Temperature Measurement, Probe Number = 0, T = 29.353000 C 12:05:28 Temperature Measurement, Probe Number = 1, T = 29.115000 C 12:05:29 Temperature Measurement, Probe Number = 2, T = 27.991000 C 12:05:29 Temperature Measurement, Probe Number = 3, T = 28.494000 C 12:05:30 Temperature Measurement, Probe Number = 4, T = 28.751000 C 12:05:30 Temperature Measurement, Probe Number = 5, T = 24.930000 C 12:05:30 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.35 P3 = RTD Outside, Value = 29.11 P4 = Supply Water, Value = 27.99 P5 = Return Water, Value = 28.49 P6 = @ RTD Outside, Value = 28.75 P7 = Ambient, Value = 24.93 12:05:31 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000020, Bz = -0.000020 T 12:05:38 Move Stage, position = -0.184000 12:05:40 Stage Position Measurement, pos = -0.1843050 m 12:05:41 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:05:41 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:05:42 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:05:43 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:05:43 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:05:43 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:05:44 Main Volt Voltage Measurement, V = 0.000002 V 12:05:45 Temperature Measurement, Probe Number = 0, T = 29.354000 C 12:05:48 Temperature Measurement, Probe Number = 1, T = 29.115000 C 12:05:49 Temperature Measurement, Probe Number = 2, T = 28.031000 C 12:05:49 Temperature Measurement, Probe Number = 3, T = 28.507000 C 12:05:50 Temperature Measurement, Probe Number = 4, T = 28.754000 C 12:05:51 Temperature Measurement, Probe Number = 5, T = 24.999000 C 12:05:51 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.35 P3 = RTD Outside, Value = 29.11 P4 = Supply Water, Value = 28.03 P5 = Return Water, Value = 28.51 P6 = @ RTD Outside, Value = 28.75 P7 = Ambient, Value = 25.00 12:05:51 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000030, Bz = -0.000030 T 12:05:58 Move Stage, position = -0.182000 12:06:00 Stage Position Measurement, pos = -0.1823150 m 12:06:01 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:06:01 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:06:02 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:06:03 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:06:03 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:06:03 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:06:04 Main Volt Voltage Measurement, V = 0.000002 V 12:06:05 Temperature Measurement, Probe Number = 0, T = 29.351000 C 12:06:08 Temperature Measurement, Probe Number = 1, T = 29.113000 C 12:06:09 Temperature Measurement, Probe Number = 2, T = 28.076000 C 12:06:10 Temperature Measurement, Probe Number = 3, T = 28.528000 C 12:06:10 Temperature Measurement, Probe Number = 4, T = 28.752000 C 12:06:11 Temperature Measurement, Probe Number = 5, T = 25.013000 C 12:06:11 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.35 P3 = RTD Outside, Value = 29.11 P4 = Supply Water, Value = 28.08 P5 = Return Water, Value = 28.53 P6 = @ RTD Outside, Value = 28.75 P7 = Ambient, Value = 25.01 12:06:11 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000020, Bz = -0.000020 T 12:06:19 Move Stage, position = -0.180000 12:06:20 Stage Position Measurement, pos = -0.1803000 m 12:06:21 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:06:22 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:06:22 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:06:23 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:06:23 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:06:24 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:06:24 Main Volt Voltage Measurement, V = 0.000001 V 12:06:25 Temperature Measurement, Probe Number = 0, T = 29.351000 C 12:06:29 Temperature Measurement, Probe Number = 1, T = 29.114000 C 12:06:29 Temperature Measurement, Probe Number = 2, T = 28.093000 C 12:06:30 Temperature Measurement, Probe Number = 3, T = 28.546000 C 12:06:30 Temperature Measurement, Probe Number = 4, T = 28.751000 C 12:06:31 Temperature Measurement, Probe Number = 5, T = 25.103000 C 12:06:31 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.35 P3 = RTD Outside, Value = 29.11 P4 = Supply Water, Value = 28.09 P5 = Return Water, Value = 28.55 P6 = @ RTD Outside, Value = 28.75 P7 = Ambient, Value = 25.10 12:06:31 3 Axis Hall Probe Field Measurement, Bx = -0.000010, By = 0.000020, Bz = -0.000030 T 12:06:39 Move Stage, position = -0.178000 12:06:40 Stage Position Measurement, pos = -0.1783100 m 12:06:41 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:06:42 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:06:42 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:06:43 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:06:43 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:06:44 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:06:44 Main Volt Voltage Measurement, V = 0.000001 V 12:06:45 Temperature Measurement, Probe Number = 0, T = 29.353000 C 12:06:49 Temperature Measurement, Probe Number = 1, T = 29.112000 C 12:06:49 Temperature Measurement, Probe Number = 2, T = 28.138000 C 12:06:50 Temperature Measurement, Probe Number = 3, T = 28.556000 C 12:06:50 Temperature Measurement, Probe Number = 4, T = 28.750000 C 12:06:51 Temperature Measurement, Probe Number = 5, T = 25.129000 C 12:06:51 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.35 P3 = RTD Outside, Value = 29.11 P4 = Supply Water, Value = 28.14 P5 = Return Water, Value = 28.56 P6 = @ RTD Outside, Value = 28.75 P7 = Ambient, Value = 25.13 12:06:51 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000030, Bz = -0.000020 T 12:06:59 Move Stage, position = -0.176000 12:07:00 Stage Position Measurement, pos = -0.1762900 m 12:07:01 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:07:02 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:07:02 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:07:03 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:07:03 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:07:04 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:07:04 Main Volt Voltage Measurement, V = 0.000001 V 12:07:05 Temperature Measurement, Probe Number = 0, T = 29.359000 C 12:07:09 Temperature Measurement, Probe Number = 1, T = 29.118000 C 12:07:09 Temperature Measurement, Probe Number = 2, T = 28.154000 C 12:07:10 Temperature Measurement, Probe Number = 3, T = 28.569000 C 12:07:10 Temperature Measurement, Probe Number = 4, T = 28.752000 C 12:07:11 Temperature Measurement, Probe Number = 5, T = 25.121000 C 12:07:11 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.36 P3 = RTD Outside, Value = 29.12 P4 = Supply Water, Value = 28.15 P5 = Return Water, Value = 28.57 P6 = @ RTD Outside, Value = 28.75 P7 = Ambient, Value = 25.12 12:07:11 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000030, Bz = -0.000020 T 12:07:19 Move Stage, position = -0.174000 12:07:20 Stage Position Measurement, pos = -0.1742550 m 12:07:21 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:07:22 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:07:23 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:07:23 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:07:23 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:07:24 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:07:25 Main Volt Voltage Measurement, V = 0.000001 V 12:07:25 Temperature Measurement, Probe Number = 0, T = 29.361000 C 12:07:29 Temperature Measurement, Probe Number = 1, T = 29.122000 C 12:07:29 Temperature Measurement, Probe Number = 2, T = 28.136000 C 12:07:30 Temperature Measurement, Probe Number = 3, T = 28.576000 C 12:07:30 Temperature Measurement, Probe Number = 4, T = 28.757000 C 12:07:31 Temperature Measurement, Probe Number = 5, T = 25.074000 C 12:07:31 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.36 P3 = RTD Outside, Value = 29.12 P4 = Supply Water, Value = 28.14 P5 = Return Water, Value = 28.58 P6 = @ RTD Outside, Value = 28.76 P7 = Ambient, Value = 25.07 12:07:31 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000020, Bz = -0.000030 T 12:07:39 Move Stage, position = -0.172000 12:07:40 Stage Position Measurement, pos = -0.1722300 m 12:07:41 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:07:42 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:07:43 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:07:43 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:07:43 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:07:44 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:07:45 Main Volt Voltage Measurement, V = 0.000001 V 12:07:45 Temperature Measurement, Probe Number = 0, T = 29.365000 C 12:07:49 Temperature Measurement, Probe Number = 1, T = 29.127000 C 12:07:49 Temperature Measurement, Probe Number = 2, T = 28.120000 C 12:07:50 Temperature Measurement, Probe Number = 3, T = 28.574000 C 12:07:51 Temperature Measurement, Probe Number = 4, T = 28.760000 C 12:07:51 Temperature Measurement, Probe Number = 5, T = 25.168000 C 12:07:51 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.36 P3 = RTD Outside, Value = 29.13 P4 = Supply Water, Value = 28.12 P5 = Return Water, Value = 28.57 P6 = @ RTD Outside, Value = 28.76 P7 = Ambient, Value = 25.17 12:07:51 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000030, Bz = -0.000020 T 12:07:59 Move Stage, position = -0.170000 12:08:01 Stage Position Measurement, pos = -0.1702200 m 12:08:01 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:08:02 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:08:03 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:08:03 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:08:03 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:08:04 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:08:05 Main Volt Voltage Measurement, V = 0.000001 V 12:08:05 Temperature Measurement, Probe Number = 0, T = 29.365000 C 12:08:09 Temperature Measurement, Probe Number = 1, T = 29.126000 C 12:08:10 Temperature Measurement, Probe Number = 2, T = 28.158000 C 12:08:10 Temperature Measurement, Probe Number = 3, T = 28.586000 C 12:08:11 Temperature Measurement, Probe Number = 4, T = 28.762000 C 12:08:11 Temperature Measurement, Probe Number = 5, T = 25.233000 C 12:08:11 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.36 P3 = RTD Outside, Value = 29.13 P4 = Supply Water, Value = 28.16 P5 = Return Water, Value = 28.59 P6 = @ RTD Outside, Value = 28.76 P7 = Ambient, Value = 25.23 12:08:11 3 Axis Hall Probe Field Measurement, Bx = -0.000010, By = 0.000030, Bz = -0.000030 T 12:08:19 Move Stage, position = -0.168000 12:08:21 Stage Position Measurement, pos = -0.1682450 m 12:08:21 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:08:22 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:08:23 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:08:24 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:08:24 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:08:24 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:08:25 Main Volt Voltage Measurement, V = 0.000001 V 12:08:25 Temperature Measurement, Probe Number = 0, T = 29.365000 C 12:08:29 Temperature Measurement, Probe Number = 1, T = 29.126000 C 12:08:30 Temperature Measurement, Probe Number = 2, T = 28.200000 C 12:08:30 Temperature Measurement, Probe Number = 3, T = 28.596000 C 12:08:31 Temperature Measurement, Probe Number = 4, T = 28.765000 C 12:08:31 Temperature Measurement, Probe Number = 5, T = 25.395000 C 12:08:31 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.36 P3 = RTD Outside, Value = 29.13 P4 = Supply Water, Value = 28.20 P5 = Return Water, Value = 28.60 P6 = @ RTD Outside, Value = 28.77 P7 = Ambient, Value = 25.39 12:08:31 3 Axis Hall Probe Field Measurement, Bx = 0.000010, By = 0.000020, Bz = -0.000020 T 12:08:39 Move Stage, position = -0.166000 12:08:41 Stage Position Measurement, pos = -0.1662400 m 12:08:42 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:08:42 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:08:43 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:08:44 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:08:44 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:08:44 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:08:45 Main Volt Voltage Measurement, V = 0.000001 V 12:08:46 Temperature Measurement, Probe Number = 0, T = 29.364000 C 12:08:49 Temperature Measurement, Probe Number = 1, T = 29.127000 C 12:08:50 Temperature Measurement, Probe Number = 2, T = 28.221000 C 12:08:50 Temperature Measurement, Probe Number = 3, T = 28.589000 C 12:08:51 Temperature Measurement, Probe Number = 4, T = 28.768000 C 12:08:52 Temperature Measurement, Probe Number = 5, T = 25.341000 C 12:08:52 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.36 P3 = RTD Outside, Value = 29.13 P4 = Supply Water, Value = 28.22 P5 = Return Water, Value = 28.59 P6 = @ RTD Outside, Value = 28.77 P7 = Ambient, Value = 25.34 12:08:52 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000020, Bz = -0.000020 T 12:08:59 Move Stage, position = -0.164000 12:09:01 Stage Position Measurement, pos = -0.1642500 m 12:09:02 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:09:02 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:09:03 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:09:04 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:09:04 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:09:05 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:09:05 Main Volt Voltage Measurement, V = 0.000001 V 12:09:06 Temperature Measurement, Probe Number = 0, T = 29.366000 C 12:09:09 Temperature Measurement, Probe Number = 1, T = 29.134000 C 12:09:10 Temperature Measurement, Probe Number = 2, T = 28.194000 C 12:09:11 Temperature Measurement, Probe Number = 3, T = 28.576000 C 12:09:11 Temperature Measurement, Probe Number = 4, T = 28.774000 C 12:09:12 Temperature Measurement, Probe Number = 5, T = 25.234000 C 12:09:12 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.37 P3 = RTD Outside, Value = 29.13 P4 = Supply Water, Value = 28.19 P5 = Return Water, Value = 28.58 P6 = @ RTD Outside, Value = 28.77 P7 = Ambient, Value = 25.23 12:09:12 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000020, Bz = -0.000030 T 12:09:19 Move Stage, position = -0.162000 12:09:21 Stage Position Measurement, pos = -0.1622300 m 12:09:22 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:09:23 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:09:23 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:09:24 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:09:24 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:09:25 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:09:25 Main Volt Voltage Measurement, V = 0.000001 V 12:09:26 Temperature Measurement, Probe Number = 0, T = 29.367000 C 12:09:29 Temperature Measurement, Probe Number = 1, T = 29.134000 C 12:09:30 Temperature Measurement, Probe Number = 2, T = 28.136000 C 12:09:31 Temperature Measurement, Probe Number = 3, T = 28.560000 C 12:09:31 Temperature Measurement, Probe Number = 4, T = 28.775000 C 12:09:32 Temperature Measurement, Probe Number = 5, T = 25.118000 C 12:09:32 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.37 P3 = RTD Outside, Value = 29.13 P4 = Supply Water, Value = 28.14 P5 = Return Water, Value = 28.56 P6 = @ RTD Outside, Value = 28.77 P7 = Ambient, Value = 25.12 12:09:32 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000020, Bz = -0.000010 T 12:09:40 Move Stage, position = -0.160000 12:09:41 Stage Position Measurement, pos = -0.1602350 m 12:09:42 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:09:43 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:09:43 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:09:44 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:09:44 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:09:45 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:09:45 Main Volt Voltage Measurement, V = 0.000001 V 12:09:46 Temperature Measurement, Probe Number = 0, T = 29.361000 C 12:09:50 Temperature Measurement, Probe Number = 1, T = 29.130000 C 12:09:50 Temperature Measurement, Probe Number = 2, T = 28.071000 C 12:09:51 Temperature Measurement, Probe Number = 3, T = 28.545000 C 12:09:51 Temperature Measurement, Probe Number = 4, T = 28.775000 C 12:09:52 Temperature Measurement, Probe Number = 5, T = 25.121000 C 12:09:52 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.36 P3 = RTD Outside, Value = 29.13 P4 = Supply Water, Value = 28.07 P5 = Return Water, Value = 28.55 P6 = @ RTD Outside, Value = 28.77 P7 = Ambient, Value = 25.12 12:09:52 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000030, Bz = -0.000010 T 12:10:00 Move Stage, position = -0.158000 12:10:01 Stage Position Measurement, pos = -0.1582250 m 12:10:02 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:10:03 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:10:03 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:10:04 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:10:04 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:10:05 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:10:05 Main Volt Voltage Measurement, V = 0.000001 V 12:10:06 Temperature Measurement, Probe Number = 0, T = 29.357000 C 12:10:10 Temperature Measurement, Probe Number = 1, T = 29.117000 C 12:10:10 Temperature Measurement, Probe Number = 2, T = 28.068000 C 12:10:11 Temperature Measurement, Probe Number = 3, T = 28.546000 C 12:10:11 Temperature Measurement, Probe Number = 4, T = 28.769000 C 12:10:12 Temperature Measurement, Probe Number = 5, T = 25.215000 C 12:10:12 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.36 P3 = RTD Outside, Value = 29.12 P4 = Supply Water, Value = 28.07 P5 = Return Water, Value = 28.55 P6 = @ RTD Outside, Value = 28.77 P7 = Ambient, Value = 25.21 12:10:12 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000020, Bz = -0.000020 T 12:10:20 Move Stage, position = -0.156000 12:10:21 Stage Position Measurement, pos = -0.1562150 m 12:10:22 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:10:23 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:10:24 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:10:24 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:10:24 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:10:25 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:10:26 Main Volt Voltage Measurement, V = 0.000001 V 12:10:26 Temperature Measurement, Probe Number = 0, T = 29.356000 C 12:10:30 Temperature Measurement, Probe Number = 1, T = 29.110000 C 12:10:30 Temperature Measurement, Probe Number = 2, T = 28.120000 C 12:10:31 Temperature Measurement, Probe Number = 3, T = 28.569000 C 12:10:32 Temperature Measurement, Probe Number = 4, T = 28.755000 C 12:10:32 Temperature Measurement, Probe Number = 5, T = 25.239000 C 12:10:32 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.36 P3 = RTD Outside, Value = 29.11 P4 = Supply Water, Value = 28.12 P5 = Return Water, Value = 28.57 P6 = @ RTD Outside, Value = 28.75 P7 = Ambient, Value = 25.24 12:10:32 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000020, Bz = -0.000020 T 12:10:40 Move Stage, position = -0.154000 12:10:42 Stage Position Measurement, pos = -0.1542000 m 12:10:42 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:10:43 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:10:44 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:10:44 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:10:44 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:10:45 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:10:46 Main Volt Voltage Measurement, V = 0.000001 V 12:10:46 Temperature Measurement, Probe Number = 0, T = 29.358000 C 12:10:50 Temperature Measurement, Probe Number = 1, T = 29.115000 C 12:10:50 Temperature Measurement, Probe Number = 2, T = 28.178000 C 12:10:51 Temperature Measurement, Probe Number = 3, T = 28.583000 C 12:10:52 Temperature Measurement, Probe Number = 4, T = 28.749000 C 12:10:52 Temperature Measurement, Probe Number = 5, T = 25.297000 C 12:10:52 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.36 P3 = RTD Outside, Value = 29.11 P4 = Supply Water, Value = 28.18 P5 = Return Water, Value = 28.58 P6 = @ RTD Outside, Value = 28.75 P7 = Ambient, Value = 25.30 12:10:52 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000010, Bz = -0.000010 T 12:11:00 Move Stage, position = -0.152000 12:11:02 Stage Position Measurement, pos = -0.1521650 m 12:11:02 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:11:03 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:11:04 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:11:04 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:11:04 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:11:05 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:11:06 Main Volt Voltage Measurement, V = 0.000001 V 12:11:06 Temperature Measurement, Probe Number = 0, T = 29.366000 C 12:11:10 Temperature Measurement, Probe Number = 1, T = 29.125000 C 12:11:11 Temperature Measurement, Probe Number = 2, T = 28.199000 C 12:11:11 Temperature Measurement, Probe Number = 3, T = 28.587000 C 12:11:12 Temperature Measurement, Probe Number = 4, T = 28.754000 C 12:11:12 Temperature Measurement, Probe Number = 5, T = 25.255000 C 12:11:12 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.37 P3 = RTD Outside, Value = 29.12 P4 = Supply Water, Value = 28.20 P5 = Return Water, Value = 28.59 P6 = @ RTD Outside, Value = 28.75 P7 = Ambient, Value = 25.25 12:11:12 3 Axis Hall Probe Field Measurement, Bx = -0.000010, By = 0.000020, Bz = -0.000020 T 12:11:20 Move Stage, position = -0.150000 12:11:22 Stage Position Measurement, pos = -0.1501850 m 12:11:22 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:11:23 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:11:24 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:11:25 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:11:25 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:11:25 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:11:26 Main Volt Voltage Measurement, V = 0.000001 V 12:11:26 Temperature Measurement, Probe Number = 0, T = 29.373000 C 12:11:30 Temperature Measurement, Probe Number = 1, T = 29.134000 C 12:11:31 Temperature Measurement, Probe Number = 2, T = 28.200000 C 12:11:31 Temperature Measurement, Probe Number = 3, T = 28.584000 C 12:11:32 Temperature Measurement, Probe Number = 4, T = 28.764000 C 12:11:32 Temperature Measurement, Probe Number = 5, T = 25.271000 C 12:11:32 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.37 P3 = RTD Outside, Value = 29.13 P4 = Supply Water, Value = 28.20 P5 = Return Water, Value = 28.58 P6 = @ RTD Outside, Value = 28.76 P7 = Ambient, Value = 25.27 12:11:32 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000030, Bz = -0.000020 T 12:11:40 Move Stage, position = -0.148000 12:11:42 Stage Position Measurement, pos = -0.1481900 m 12:11:43 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:11:43 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:11:44 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:11:45 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:11:45 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:11:45 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:11:46 Main Volt Voltage Measurement, V = 0.000001 V 12:11:47 Temperature Measurement, Probe Number = 0, T = 29.376000 C 12:11:50 Temperature Measurement, Probe Number = 1, T = 29.136000 C 12:11:51 Temperature Measurement, Probe Number = 2, T = 28.189000 C 12:11:51 Temperature Measurement, Probe Number = 3, T = 28.586000 C 12:11:52 Temperature Measurement, Probe Number = 4, T = 28.772000 C 12:11:53 Temperature Measurement, Probe Number = 5, T = 25.228000 C 12:11:53 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.38 P3 = RTD Outside, Value = 29.14 P4 = Supply Water, Value = 28.19 P5 = Return Water, Value = 28.59 P6 = @ RTD Outside, Value = 28.77 P7 = Ambient, Value = 25.23 12:11:53 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000020, Bz = -0.000020 T 12:12:00 Move Stage, position = -0.146000 12:12:02 Stage Position Measurement, pos = -0.1462150 m 12:12:03 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:12:03 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:12:04 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:12:05 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:12:05 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:12:05 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:12:06 Main Volt Voltage Measurement, V = 0.000001 V 12:12:07 Temperature Measurement, Probe Number = 0, T = 29.375000 C 12:12:10 Temperature Measurement, Probe Number = 1, T = 29.138000 C 12:12:11 Temperature Measurement, Probe Number = 2, T = 28.200000 C 12:12:11 Temperature Measurement, Probe Number = 3, T = 28.585000 C 12:12:12 Temperature Measurement, Probe Number = 4, T = 28.778000 C 12:12:13 Temperature Measurement, Probe Number = 5, T = 25.194000 C 12:12:13 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.38 P3 = RTD Outside, Value = 29.14 P4 = Supply Water, Value = 28.20 P5 = Return Water, Value = 28.59 P6 = @ RTD Outside, Value = 28.78 P7 = Ambient, Value = 25.19 12:12:13 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000030, Bz = -0.000030 T 12:12:20 Move Stage, position = -0.144000 12:12:22 Stage Position Measurement, pos = -0.1441800 m 12:12:23 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:12:23 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:12:24 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:12:25 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:12:25 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:12:26 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:12:26 Main Volt Voltage Measurement, V = 0.000001 V 12:12:27 Temperature Measurement, Probe Number = 0, T = 29.372000 C 12:12:31 Temperature Measurement, Probe Number = 1, T = 29.135000 C 12:12:31 Temperature Measurement, Probe Number = 2, T = 28.188000 C 12:12:32 Temperature Measurement, Probe Number = 3, T = 28.593000 C 12:12:32 Temperature Measurement, Probe Number = 4, T = 28.781000 C 12:12:33 Temperature Measurement, Probe Number = 5, T = 25.177000 C 12:12:33 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.37 P3 = RTD Outside, Value = 29.14 P4 = Supply Water, Value = 28.19 P5 = Return Water, Value = 28.59 P6 = @ RTD Outside, Value = 28.78 P7 = Ambient, Value = 25.18 12:12:33 3 Axis Hall Probe Field Measurement, Bx = -0.000010, By = 0.000020, Bz = -0.000020 T 12:12:41 Move Stage, position = -0.142000 12:12:42 Stage Position Measurement, pos = -0.1421850 m 12:12:43 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:12:44 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:12:44 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:12:45 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:12:45 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:12:46 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:12:46 Main Volt Voltage Measurement, V = 0.000001 V 12:12:47 Temperature Measurement, Probe Number = 0, T = 29.372000 C 12:12:51 Temperature Measurement, Probe Number = 1, T = 29.137000 C 12:12:51 Temperature Measurement, Probe Number = 2, T = 28.188000 C 12:12:52 Temperature Measurement, Probe Number = 3, T = 28.598000 C 12:12:52 Temperature Measurement, Probe Number = 4, T = 28.783000 C 12:12:53 Temperature Measurement, Probe Number = 5, T = 25.150000 C 12:12:53 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.37 P3 = RTD Outside, Value = 29.14 P4 = Supply Water, Value = 28.19 P5 = Return Water, Value = 28.60 P6 = @ RTD Outside, Value = 28.78 P7 = Ambient, Value = 25.15 12:12:53 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000020, Bz = -0.000030 T 12:13:01 Move Stage, position = -0.140000 12:13:02 Stage Position Measurement, pos = -0.1401700 m 12:13:03 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:13:04 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:13:05 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:13:05 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:13:05 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:13:06 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:13:07 Main Volt Voltage Measurement, V = 0.000001 V 12:13:07 Temperature Measurement, Probe Number = 0, T = 29.369000 C 12:13:11 Temperature Measurement, Probe Number = 1, T = 29.128000 C 12:13:11 Temperature Measurement, Probe Number = 2, T = 28.186000 C 12:13:12 Temperature Measurement, Probe Number = 3, T = 28.577000 C 12:13:13 Temperature Measurement, Probe Number = 4, T = 28.784000 C 12:13:13 Temperature Measurement, Probe Number = 5, T = 25.067000 C 12:13:13 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.37 P3 = RTD Outside, Value = 29.13 P4 = Supply Water, Value = 28.19 P5 = Return Water, Value = 28.58 P6 = @ RTD Outside, Value = 28.78 P7 = Ambient, Value = 25.07 12:13:13 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000020, Bz = -0.000010 T 12:13:21 Move Stage, position = -0.138000 12:13:23 Stage Position Measurement, pos = -0.1381600 m 12:13:23 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:13:24 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:13:25 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:13:25 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:13:25 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:13:26 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:13:27 Main Volt Voltage Measurement, V = 0.000002 V 12:13:27 Temperature Measurement, Probe Number = 0, T = 29.363000 C 12:13:31 Temperature Measurement, Probe Number = 1, T = 29.129000 C 12:13:32 Temperature Measurement, Probe Number = 2, T = 28.149000 C 12:13:32 Temperature Measurement, Probe Number = 3, T = 28.543000 C 12:13:33 Temperature Measurement, Probe Number = 4, T = 28.782000 C 12:13:33 Temperature Measurement, Probe Number = 5, T = 25.035000 C 12:13:33 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.36 P3 = RTD Outside, Value = 29.13 P4 = Supply Water, Value = 28.15 P5 = Return Water, Value = 28.54 P6 = @ RTD Outside, Value = 28.78 P7 = Ambient, Value = 25.04 12:13:33 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000010, Bz = -0.000020 T 12:13:41 Move Stage, position = -0.136000 12:13:43 Stage Position Measurement, pos = -0.1361550 m 12:13:43 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:13:44 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:13:45 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:13:46 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:13:46 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:13:46 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:13:47 Main Volt Voltage Measurement, V = 0.000001 V 12:13:47 Temperature Measurement, Probe Number = 0, T = 29.360000 C 12:13:51 Temperature Measurement, Probe Number = 1, T = 29.126000 C 12:13:52 Temperature Measurement, Probe Number = 2, T = 28.137000 C 12:13:52 Temperature Measurement, Probe Number = 3, T = 28.481000 C 12:13:53 Temperature Measurement, Probe Number = 4, T = 28.783000 C 12:13:53 Temperature Measurement, Probe Number = 5, T = 24.958000 C 12:13:53 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.36 P3 = RTD Outside, Value = 29.13 P4 = Supply Water, Value = 28.14 P5 = Return Water, Value = 28.48 P6 = @ RTD Outside, Value = 28.78 P7 = Ambient, Value = 24.96 12:13:53 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000020, Bz = -0.000020 T 12:14:01 Move Stage, position = -0.134000 12:14:03 Stage Position Measurement, pos = -0.1341300 m 12:14:03 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:14:04 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:14:05 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:14:06 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:14:06 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:14:06 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:14:07 Main Volt Voltage Measurement, V = 0.000002 V 12:14:07 Temperature Measurement, Probe Number = 0, T = 29.356000 C 12:14:11 Temperature Measurement, Probe Number = 1, T = 29.120000 C 12:14:12 Temperature Measurement, Probe Number = 2, T = 28.139000 C 12:14:12 Temperature Measurement, Probe Number = 3, T = 28.426000 C 12:14:13 Temperature Measurement, Probe Number = 4, T = 28.782000 C 12:14:13 Temperature Measurement, Probe Number = 5, T = 24.889000 C 12:14:13 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.36 P3 = RTD Outside, Value = 29.12 P4 = Supply Water, Value = 28.14 P5 = Return Water, Value = 28.43 P6 = @ RTD Outside, Value = 28.78 P7 = Ambient, Value = 24.89 12:14:14 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000020, Bz = -0.000020 T 12:14:21 Move Stage, position = -0.132000 12:14:23 Stage Position Measurement, pos = -0.1321450 m 12:14:24 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:14:24 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:14:25 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:14:26 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:14:26 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:14:26 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:14:27 Main Volt Voltage Measurement, V = 0.000002 V 12:14:28 Temperature Measurement, Probe Number = 0, T = 29.351000 C 12:14:31 Temperature Measurement, Probe Number = 1, T = 29.113000 C 12:14:32 Temperature Measurement, Probe Number = 2, T = 28.110000 C 12:14:32 Temperature Measurement, Probe Number = 3, T = 28.413000 C 12:14:33 Temperature Measurement, Probe Number = 4, T = 28.775000 C 12:14:34 Temperature Measurement, Probe Number = 5, T = 24.864000 C 12:14:34 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.35 P3 = RTD Outside, Value = 29.11 P4 = Supply Water, Value = 28.11 P5 = Return Water, Value = 28.41 P6 = @ RTD Outside, Value = 28.77 P7 = Ambient, Value = 24.86 12:14:34 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000030, Bz = -0.000030 T 12:14:41 Move Stage, position = -0.130000 12:14:43 Stage Position Measurement, pos = -0.1301400 m 12:14:44 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:14:44 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:14:45 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:14:46 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:14:46 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:14:47 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:14:47 Main Volt Voltage Measurement, V = 0.000002 V 12:14:48 Temperature Measurement, Probe Number = 0, T = 29.344000 C 12:14:51 Temperature Measurement, Probe Number = 1, T = 29.110000 C 12:14:52 Temperature Measurement, Probe Number = 2, T = 28.065000 C 12:14:52 Temperature Measurement, Probe Number = 3, T = 28.420000 C 12:14:53 Temperature Measurement, Probe Number = 4, T = 28.770000 C 12:14:54 Temperature Measurement, Probe Number = 5, T = 24.772000 C 12:14:54 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.34 P3 = RTD Outside, Value = 29.11 P4 = Supply Water, Value = 28.07 P5 = Return Water, Value = 28.42 P6 = @ RTD Outside, Value = 28.77 P7 = Ambient, Value = 24.77 12:14:54 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000010, Bz = -0.000020 T 12:15:01 Move Stage, position = -0.128000 12:15:03 Stage Position Measurement, pos = -0.1281750 m 12:15:04 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:15:04 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:15:05 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:15:06 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:15:06 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:15:07 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:15:07 Main Volt Voltage Measurement, V = 0.000002 V 12:15:08 Temperature Measurement, Probe Number = 0, T = 29.338000 C 12:15:11 Temperature Measurement, Probe Number = 1, T = 29.095000 C 12:15:12 Temperature Measurement, Probe Number = 2, T = 28.028000 C 12:15:13 Temperature Measurement, Probe Number = 3, T = 28.372000 C 12:15:13 Temperature Measurement, Probe Number = 4, T = 28.762000 C 12:15:14 Temperature Measurement, Probe Number = 5, T = 24.806000 C 12:15:14 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.34 P3 = RTD Outside, Value = 29.09 P4 = Supply Water, Value = 28.03 P5 = Return Water, Value = 28.37 P6 = @ RTD Outside, Value = 28.76 P7 = Ambient, Value = 24.81 12:15:14 3 Axis Hall Probe Field Measurement, Bx = -0.000010, By = 0.000020, Bz = 0.000000 T 12:15:22 Move Stage, position = -0.126000 12:15:23 Stage Position Measurement, pos = -0.1261750 m 12:15:24 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:15:25 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:15:25 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:15:26 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:15:26 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:15:27 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:15:27 Main Volt Voltage Measurement, V = 0.000002 V 12:15:28 Temperature Measurement, Probe Number = 0, T = 29.338000 C 12:15:32 Temperature Measurement, Probe Number = 1, T = 29.094000 C 12:15:32 Temperature Measurement, Probe Number = 2, T = 28.028000 C 12:15:33 Temperature Measurement, Probe Number = 3, T = 28.355000 C 12:15:33 Temperature Measurement, Probe Number = 4, T = 28.758000 C 12:15:34 Temperature Measurement, Probe Number = 5, T = 24.884000 C 12:15:34 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.34 P3 = RTD Outside, Value = 29.09 P4 = Supply Water, Value = 28.03 P5 = Return Water, Value = 28.36 P6 = @ RTD Outside, Value = 28.76 P7 = Ambient, Value = 24.88 12:15:34 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000010, Bz = -0.000020 T 12:15:42 Move Stage, position = -0.124000 12:15:43 Stage Position Measurement, pos = -0.1241700 m 12:15:44 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:15:45 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:15:45 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:15:46 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:15:46 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:15:47 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:15:47 Main Volt Voltage Measurement, V = 0.000002 V 12:15:48 Temperature Measurement, Probe Number = 0, T = 29.336000 C 12:15:52 Temperature Measurement, Probe Number = 1, T = 29.096000 C 12:15:52 Temperature Measurement, Probe Number = 2, T = 28.055000 C 12:15:53 Temperature Measurement, Probe Number = 3, T = 28.352000 C 12:15:53 Temperature Measurement, Probe Number = 4, T = 28.754000 C 12:15:54 Temperature Measurement, Probe Number = 5, T = 24.803000 C 12:15:54 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.34 P3 = RTD Outside, Value = 29.10 P4 = Supply Water, Value = 28.05 P5 = Return Water, Value = 28.35 P6 = @ RTD Outside, Value = 28.75 P7 = Ambient, Value = 24.80 12:15:54 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000020, Bz = -0.000020 T 12:16:02 Move Stage, position = -0.122000 12:16:03 Stage Position Measurement, pos = -0.1221500 m 12:16:04 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:16:05 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:16:06 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:16:06 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:16:06 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:16:07 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:16:08 Main Volt Voltage Measurement, V = 0.000002 V 12:16:08 Temperature Measurement, Probe Number = 0, T = 29.331000 C 12:16:12 Temperature Measurement, Probe Number = 1, T = 29.092000 C 12:16:12 Temperature Measurement, Probe Number = 2, T = 28.065000 C 12:16:13 Temperature Measurement, Probe Number = 3, T = 28.337000 C 12:16:14 Temperature Measurement, Probe Number = 4, T = 28.751000 C 12:16:14 Temperature Measurement, Probe Number = 5, T = 24.724000 C 12:16:14 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.33 P3 = RTD Outside, Value = 29.09 P4 = Supply Water, Value = 28.07 P5 = Return Water, Value = 28.34 P6 = @ RTD Outside, Value = 28.75 P7 = Ambient, Value = 24.72 12:16:14 3 Axis Hall Probe Field Measurement, Bx = -0.000010, By = 0.000010, Bz = -0.000010 T 12:16:22 Move Stage, position = -0.120000 12:16:24 Stage Position Measurement, pos = -0.1201350 m 12:16:24 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:16:25 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:16:26 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:16:27 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:16:27 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:16:27 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:16:28 Main Volt Voltage Measurement, V = 0.000002 V 12:16:28 Temperature Measurement, Probe Number = 0, T = 29.327000 C 12:16:32 Temperature Measurement, Probe Number = 1, T = 29.089000 C 12:16:33 Temperature Measurement, Probe Number = 2, T = 28.065000 C 12:16:33 Temperature Measurement, Probe Number = 3, T = 28.314000 C 12:16:34 Temperature Measurement, Probe Number = 4, T = 28.746000 C 12:16:34 Temperature Measurement, Probe Number = 5, T = 24.729000 C 12:16:34 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.33 P3 = RTD Outside, Value = 29.09 P4 = Supply Water, Value = 28.07 P5 = Return Water, Value = 28.31 P6 = @ RTD Outside, Value = 28.75 P7 = Ambient, Value = 24.73 12:16:34 3 Axis Hall Probe Field Measurement, Bx = -0.000010, By = 0.000010, Bz = 0.000000 T 12:16:42 Move Stage, position = -0.118000 12:16:44 Stage Position Measurement, pos = -0.1181300 m 12:16:44 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:16:45 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:16:46 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:16:47 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:16:47 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:16:47 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:16:48 Main Volt Voltage Measurement, V = 0.000003 V 12:16:49 Temperature Measurement, Probe Number = 0, T = 29.324000 C 12:16:52 Temperature Measurement, Probe Number = 1, T = 29.086000 C 12:16:53 Temperature Measurement, Probe Number = 2, T = 28.063000 C 12:16:53 Temperature Measurement, Probe Number = 3, T = 28.318000 C 12:16:54 Temperature Measurement, Probe Number = 4, T = 28.742000 C 12:16:54 Temperature Measurement, Probe Number = 5, T = 24.724000 C 12:16:54 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.32 P3 = RTD Outside, Value = 29.09 P4 = Supply Water, Value = 28.06 P5 = Return Water, Value = 28.32 P6 = @ RTD Outside, Value = 28.74 P7 = Ambient, Value = 24.72 12:16:55 3 Axis Hall Probe Field Measurement, Bx = -0.000010, By = 0.000010, Bz = -0.000010 T 12:17:02 Move Stage, position = -0.116000 12:17:04 Stage Position Measurement, pos = -0.1161250 m 12:17:05 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:17:05 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:17:06 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:17:07 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:17:07 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:17:07 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:17:08 Main Volt Voltage Measurement, V = 0.000002 V 12:17:09 Temperature Measurement, Probe Number = 0, T = 29.323000 C 12:17:12 Temperature Measurement, Probe Number = 1, T = 29.084000 C 12:17:13 Temperature Measurement, Probe Number = 2, T = 28.080000 C 12:17:13 Temperature Measurement, Probe Number = 3, T = 28.330000 C 12:17:14 Temperature Measurement, Probe Number = 4, T = 28.739000 C 12:17:15 Temperature Measurement, Probe Number = 5, T = 24.753000 C 12:17:15 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.32 P3 = RTD Outside, Value = 29.08 P4 = Supply Water, Value = 28.08 P5 = Return Water, Value = 28.33 P6 = @ RTD Outside, Value = 28.74 P7 = Ambient, Value = 24.75 12:17:15 3 Axis Hall Probe Field Measurement, Bx = -0.000010, By = 0.000020, Bz = -0.000010 T 12:17:22 Move Stage, position = -0.114000 12:17:24 Stage Position Measurement, pos = -0.1141350 m 12:17:25 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:17:25 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:17:26 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:17:27 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:17:27 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:17:28 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:17:28 Main Volt Voltage Measurement, V = 0.000003 V 12:17:29 Temperature Measurement, Probe Number = 0, T = 29.327000 C 12:17:32 Temperature Measurement, Probe Number = 1, T = 29.085000 C 12:17:33 Temperature Measurement, Probe Number = 2, T = 28.122000 C 12:17:34 Temperature Measurement, Probe Number = 3, T = 28.364000 C 12:17:34 Temperature Measurement, Probe Number = 4, T = 28.736000 C 12:17:35 Temperature Measurement, Probe Number = 5, T = 24.824000 C 12:17:35 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.33 P3 = RTD Outside, Value = 29.09 P4 = Supply Water, Value = 28.12 P5 = Return Water, Value = 28.36 P6 = @ RTD Outside, Value = 28.74 P7 = Ambient, Value = 24.82 12:17:35 3 Axis Hall Probe Field Measurement, Bx = -0.000010, By = 0.000010, Bz = -0.000020 T 12:17:42 Move Stage, position = -0.112000 12:17:44 Stage Position Measurement, pos = -0.1121250 m 12:17:45 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:17:45 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:17:46 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:17:47 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:17:47 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:17:48 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:17:48 Main Volt Voltage Measurement, V = 0.000002 V 12:17:49 Temperature Measurement, Probe Number = 0, T = 29.331000 C 12:17:52 Temperature Measurement, Probe Number = 1, T = 29.083000 C 12:17:53 Temperature Measurement, Probe Number = 2, T = 28.170000 C 12:17:54 Temperature Measurement, Probe Number = 3, T = 28.417000 C 12:17:54 Temperature Measurement, Probe Number = 4, T = 28.731000 C 12:17:55 Temperature Measurement, Probe Number = 5, T = 25.042000 C 12:17:55 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.33 P3 = RTD Outside, Value = 29.08 P4 = Supply Water, Value = 28.17 P5 = Return Water, Value = 28.42 P6 = @ RTD Outside, Value = 28.73 P7 = Ambient, Value = 25.04 12:17:55 3 Axis Hall Probe Field Measurement, Bx = -0.000010, By = 0.000010, Bz = -0.000010 T 12:18:03 Move Stage, position = -0.110000 12:18:04 Stage Position Measurement, pos = -0.1101500 m 12:18:05 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:18:06 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:18:06 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:18:07 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:18:07 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:18:08 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:18:08 Main Volt Voltage Measurement, V = 0.000002 V 12:18:09 Temperature Measurement, Probe Number = 0, T = 29.333000 C 12:18:13 Temperature Measurement, Probe Number = 1, T = 29.066000 C 12:18:13 Temperature Measurement, Probe Number = 2, T = 28.159000 C 12:18:14 Temperature Measurement, Probe Number = 3, T = 28.464000 C 12:18:14 Temperature Measurement, Probe Number = 4, T = 28.702000 C 12:18:15 Temperature Measurement, Probe Number = 5, T = 25.037000 C 12:18:15 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.33 P3 = RTD Outside, Value = 29.07 P4 = Supply Water, Value = 28.16 P5 = Return Water, Value = 28.46 P6 = @ RTD Outside, Value = 28.70 P7 = Ambient, Value = 25.04 12:18:15 3 Axis Hall Probe Field Measurement, Bx = -0.000010, By = 0.000010, Bz = -0.000010 T 12:18:23 Move Stage, position = -0.108000 12:18:24 Stage Position Measurement, pos = -0.1081450 m 12:18:25 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:18:26 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:18:26 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:18:27 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:18:27 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:18:28 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:18:28 Main Volt Voltage Measurement, V = 0.000002 V 12:18:29 Temperature Measurement, Probe Number = 0, T = 29.341000 C 12:18:33 Temperature Measurement, Probe Number = 1, T = 29.069000 C 12:18:33 Temperature Measurement, Probe Number = 2, T = 28.122000 C 12:18:34 Temperature Measurement, Probe Number = 3, T = 28.477000 C 12:18:34 Temperature Measurement, Probe Number = 4, T = 28.677000 C 12:18:35 Temperature Measurement, Probe Number = 5, T = 25.013000 C 12:18:35 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.34 P3 = RTD Outside, Value = 29.07 P4 = Supply Water, Value = 28.12 P5 = Return Water, Value = 28.48 P6 = @ RTD Outside, Value = 28.68 P7 = Ambient, Value = 25.01 12:18:35 3 Axis Hall Probe Field Measurement, Bx = -0.000010, By = 0.000020, Bz = -0.000010 T 12:18:43 Move Stage, position = -0.106000 12:18:44 Stage Position Measurement, pos = -0.1061600 m 12:18:45 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:18:46 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:18:47 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:18:47 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:18:47 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:18:48 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:18:49 Main Volt Voltage Measurement, V = 0.000002 V 12:18:49 Temperature Measurement, Probe Number = 0, T = 29.348000 C 12:18:53 Temperature Measurement, Probe Number = 1, T = 29.088000 C 12:18:53 Temperature Measurement, Probe Number = 2, T = 28.130000 C 12:18:54 Temperature Measurement, Probe Number = 3, T = 28.482000 C 12:18:55 Temperature Measurement, Probe Number = 4, T = 28.689000 C 12:18:55 Temperature Measurement, Probe Number = 5, T = 25.070000 C 12:18:55 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.35 P3 = RTD Outside, Value = 29.09 P4 = Supply Water, Value = 28.13 P5 = Return Water, Value = 28.48 P6 = @ RTD Outside, Value = 28.69 P7 = Ambient, Value = 25.07 12:18:55 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000020, Bz = -0.000010 T 12:19:03 Move Stage, position = -0.104000 12:19:04 Stage Position Measurement, pos = -0.1041400 m 12:19:05 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:19:06 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:19:07 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:19:07 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:19:07 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:19:08 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:19:09 Main Volt Voltage Measurement, V = 0.000002 V 12:19:09 Temperature Measurement, Probe Number = 0, T = 29.353000 C 12:19:13 Temperature Measurement, Probe Number = 1, T = 29.098000 C 12:19:13 Temperature Measurement, Probe Number = 2, T = 28.167000 C 12:19:14 Temperature Measurement, Probe Number = 3, T = 28.502000 C 12:19:15 Temperature Measurement, Probe Number = 4, T = 28.703000 C 12:19:15 Temperature Measurement, Probe Number = 5, T = 25.107000 C 12:19:15 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.35 P3 = RTD Outside, Value = 29.10 P4 = Supply Water, Value = 28.17 P5 = Return Water, Value = 28.50 P6 = @ RTD Outside, Value = 28.70 P7 = Ambient, Value = 25.11 12:19:15 3 Axis Hall Probe Field Measurement, Bx = -0.000010, By = 0.000010, Bz = -0.000020 T 12:19:23 Move Stage, position = -0.102000 12:19:25 Stage Position Measurement, pos = -0.1021150 m 12:19:25 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:19:26 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:19:27 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:19:27 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:19:27 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:19:28 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:19:29 Main Volt Voltage Measurement, V = 0.000002 V 12:19:29 Temperature Measurement, Probe Number = 0, T = 29.354000 C 12:19:33 Temperature Measurement, Probe Number = 1, T = 29.105000 C 12:19:34 Temperature Measurement, Probe Number = 2, T = 28.221000 C 12:19:34 Temperature Measurement, Probe Number = 3, T = 28.533000 C 12:19:35 Temperature Measurement, Probe Number = 4, T = 28.714000 C 12:19:35 Temperature Measurement, Probe Number = 5, T = 25.155000 C 12:19:35 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.35 P3 = RTD Outside, Value = 29.11 P4 = Supply Water, Value = 28.22 P5 = Return Water, Value = 28.53 P6 = @ RTD Outside, Value = 28.71 P7 = Ambient, Value = 25.16 12:19:35 3 Axis Hall Probe Field Measurement, Bx = -0.000010, By = 0.000010, Bz = -0.000010 T 12:19:43 Move Stage, position = -0.100000 12:19:45 Stage Position Measurement, pos = -0.1001050 m 12:19:45 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:19:46 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:19:47 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:19:48 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:19:48 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:19:48 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:19:49 Main Volt Voltage Measurement, V = 0.000002 V 12:19:49 Temperature Measurement, Probe Number = 0, T = 29.357000 C 12:19:53 Temperature Measurement, Probe Number = 1, T = 29.113000 C 12:19:54 Temperature Measurement, Probe Number = 2, T = 28.230000 C 12:19:54 Temperature Measurement, Probe Number = 3, T = 28.554000 C 12:19:55 Temperature Measurement, Probe Number = 4, T = 28.725000 C 12:19:55 Temperature Measurement, Probe Number = 5, T = 25.141000 C 12:19:55 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.36 P3 = RTD Outside, Value = 29.11 P4 = Supply Water, Value = 28.23 P5 = Return Water, Value = 28.55 P6 = @ RTD Outside, Value = 28.73 P7 = Ambient, Value = 25.14 12:19:55 3 Axis Hall Probe Field Measurement, Bx = -0.000010, By = 0.000010, Bz = -0.000020 T 12:20:03 Move Stage, position = -0.098000 12:20:05 Stage Position Measurement, pos = -0.0981050 m 12:20:06 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:20:06 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:20:07 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:20:08 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:20:08 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:20:08 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:20:09 Main Volt Voltage Measurement, V = 0.000002 V 12:20:10 Temperature Measurement, Probe Number = 0, T = 29.360000 C 12:20:13 Temperature Measurement, Probe Number = 1, T = 29.114000 C 12:20:14 Temperature Measurement, Probe Number = 2, T = 28.227000 C 12:20:14 Temperature Measurement, Probe Number = 3, T = 28.576000 C 12:20:15 Temperature Measurement, Probe Number = 4, T = 28.735000 C 12:20:16 Temperature Measurement, Probe Number = 5, T = 25.154000 C 12:20:16 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.36 P3 = RTD Outside, Value = 29.11 P4 = Supply Water, Value = 28.23 P5 = Return Water, Value = 28.58 P6 = @ RTD Outside, Value = 28.73 P7 = Ambient, Value = 25.15 12:20:16 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000010, Bz = -0.000020 T 12:20:23 Move Stage, position = -0.096000 12:20:25 Stage Position Measurement, pos = -0.0961350 m 12:20:26 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:20:26 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:20:27 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:20:28 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:20:28 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:20:28 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:20:29 Main Volt Voltage Measurement, V = 0.000002 V 12:20:30 Temperature Measurement, Probe Number = 0, T = 29.361000 C 12:20:33 Temperature Measurement, Probe Number = 1, T = 29.115000 C 12:20:34 Temperature Measurement, Probe Number = 2, T = 28.228000 C 12:20:35 Temperature Measurement, Probe Number = 3, T = 28.598000 C 12:20:35 Temperature Measurement, Probe Number = 4, T = 28.742000 C 12:20:36 Temperature Measurement, Probe Number = 5, T = 25.224000 C 12:20:36 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.36 P3 = RTD Outside, Value = 29.11 P4 = Supply Water, Value = 28.23 P5 = Return Water, Value = 28.60 P6 = @ RTD Outside, Value = 28.74 P7 = Ambient, Value = 25.22 12:20:36 3 Axis Hall Probe Field Measurement, Bx = -0.000010, By = 0.000020, Bz = -0.000020 T 12:20:44 Move Stage, position = -0.094000 12:20:45 Stage Position Measurement, pos = -0.0941250 m 12:20:46 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:20:47 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:20:47 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:20:48 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:20:48 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:20:49 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:20:49 Main Volt Voltage Measurement, V = 0.000001 V 12:20:50 Temperature Measurement, Probe Number = 0, T = 29.360000 C 12:20:54 Temperature Measurement, Probe Number = 1, T = 29.111000 C 12:20:54 Temperature Measurement, Probe Number = 2, T = 28.192000 C 12:20:55 Temperature Measurement, Probe Number = 3, T = 28.600000 C 12:20:55 Temperature Measurement, Probe Number = 4, T = 28.740000 C 12:20:56 Temperature Measurement, Probe Number = 5, T = 25.190000 C 12:20:56 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.36 P3 = RTD Outside, Value = 29.11 P4 = Supply Water, Value = 28.19 P5 = Return Water, Value = 28.60 P6 = @ RTD Outside, Value = 28.74 P7 = Ambient, Value = 25.19 12:20:56 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000010, Bz = -0.000020 T 12:21:04 Move Stage, position = -0.092000 12:21:05 Stage Position Measurement, pos = -0.0921450 m 12:21:06 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:21:07 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:21:07 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:21:08 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:21:08 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:21:09 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:21:09 Main Volt Voltage Measurement, V = 0.000002 V 12:21:10 Temperature Measurement, Probe Number = 0, T = 29.362000 C 12:21:14 Temperature Measurement, Probe Number = 1, T = 29.121000 C 12:21:14 Temperature Measurement, Probe Number = 2, T = 28.156000 C 12:21:15 Temperature Measurement, Probe Number = 3, T = 28.574000 C 12:21:15 Temperature Measurement, Probe Number = 4, T = 28.749000 C 12:21:16 Temperature Measurement, Probe Number = 5, T = 25.172000 C 12:21:16 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.36 P3 = RTD Outside, Value = 29.12 P4 = Supply Water, Value = 28.16 P5 = Return Water, Value = 28.57 P6 = @ RTD Outside, Value = 28.75 P7 = Ambient, Value = 25.17 12:21:16 3 Axis Hall Probe Field Measurement, Bx = -0.000010, By = 0.000020, Bz = -0.000020 T 12:21:24 Move Stage, position = -0.090000 12:21:25 Stage Position Measurement, pos = -0.0901400 m 12:21:26 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:21:27 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:21:28 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:21:28 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:21:28 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:21:29 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:21:30 Main Volt Voltage Measurement, V = 0.000002 V 12:21:30 Temperature Measurement, Probe Number = 0, T = 29.364000 C 12:21:34 Temperature Measurement, Probe Number = 1, T = 29.123000 C 12:21:34 Temperature Measurement, Probe Number = 2, T = 28.144000 C 12:21:35 Temperature Measurement, Probe Number = 3, T = 28.577000 C 12:21:36 Temperature Measurement, Probe Number = 4, T = 28.755000 C 12:21:36 Temperature Measurement, Probe Number = 5, T = 25.175000 C 12:21:36 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.36 P3 = RTD Outside, Value = 29.12 P4 = Supply Water, Value = 28.14 P5 = Return Water, Value = 28.58 P6 = @ RTD Outside, Value = 28.75 P7 = Ambient, Value = 25.18 12:21:36 3 Axis Hall Probe Field Measurement, Bx = 0.000010, By = 0.000010, Bz = -0.000030 T 12:21:44 Move Stage, position = -0.088000 12:21:46 Stage Position Measurement, pos = -0.0881450 m 12:21:46 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:21:47 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:21:48 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:21:48 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:21:48 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:21:49 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:21:50 Main Volt Voltage Measurement, V = 0.000002 V 12:21:50 Temperature Measurement, Probe Number = 0, T = 29.363000 C 12:21:54 Temperature Measurement, Probe Number = 1, T = 29.127000 C 12:21:54 Temperature Measurement, Probe Number = 2, T = 28.146000 C 12:21:55 Temperature Measurement, Probe Number = 3, T = 28.586000 C 12:21:56 Temperature Measurement, Probe Number = 4, T = 28.764000 C 12:21:56 Temperature Measurement, Probe Number = 5, T = 25.193000 C 12:21:56 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.36 P3 = RTD Outside, Value = 29.13 P4 = Supply Water, Value = 28.15 P5 = Return Water, Value = 28.59 P6 = @ RTD Outside, Value = 28.76 P7 = Ambient, Value = 25.19 12:21:56 3 Axis Hall Probe Field Measurement, Bx = -0.000010, By = 0.000020, Bz = -0.000030 T 12:22:04 Move Stage, position = -0.086000 12:22:06 Stage Position Measurement, pos = -0.0861400 m 12:22:06 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:22:07 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:22:08 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:22:08 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:22:08 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:22:09 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:22:10 Main Volt Voltage Measurement, V = 0.000002 V 12:22:10 Temperature Measurement, Probe Number = 0, T = 29.366000 C 12:22:14 Temperature Measurement, Probe Number = 1, T = 29.128000 C 12:22:15 Temperature Measurement, Probe Number = 2, T = 28.124000 C 12:22:15 Temperature Measurement, Probe Number = 3, T = 28.580000 C 12:22:16 Temperature Measurement, Probe Number = 4, T = 28.770000 C 12:22:16 Temperature Measurement, Probe Number = 5, T = 25.211000 C 12:22:16 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.37 P3 = RTD Outside, Value = 29.13 P4 = Supply Water, Value = 28.12 P5 = Return Water, Value = 28.58 P6 = @ RTD Outside, Value = 28.77 P7 = Ambient, Value = 25.21 12:22:16 3 Axis Hall Probe Field Measurement, Bx = 0.000010, By = 0.000020, Bz = -0.000040 T 12:22:24 Move Stage, position = -0.084000 12:22:26 Stage Position Measurement, pos = -0.0841250 m 12:22:26 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:22:27 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:22:28 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:22:29 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:22:29 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:22:29 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:22:30 Main Volt Voltage Measurement, V = 0.000001 V 12:22:30 Temperature Measurement, Probe Number = 0, T = 29.368000 C 12:22:34 Temperature Measurement, Probe Number = 1, T = 29.130000 C 12:22:35 Temperature Measurement, Probe Number = 2, T = 28.107000 C 12:22:35 Temperature Measurement, Probe Number = 3, T = 28.583000 C 12:22:36 Temperature Measurement, Probe Number = 4, T = 28.773000 C 12:22:36 Temperature Measurement, Probe Number = 5, T = 25.121000 C 12:22:36 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.37 P3 = RTD Outside, Value = 29.13 P4 = Supply Water, Value = 28.11 P5 = Return Water, Value = 28.58 P6 = @ RTD Outside, Value = 28.77 P7 = Ambient, Value = 25.12 12:22:37 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000010, Bz = -0.000050 T 12:22:44 Move Stage, position = -0.082000 12:22:46 Stage Position Measurement, pos = -0.0821050 m 12:22:47 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:22:47 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:22:48 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:22:49 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:22:49 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:22:50 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:22:50 Main Volt Voltage Measurement, V = 0.000002 V 12:22:51 Temperature Measurement, Probe Number = 0, T = 29.369000 C 12:22:54 Temperature Measurement, Probe Number = 1, T = 29.133000 C 12:22:55 Temperature Measurement, Probe Number = 2, T = 28.116000 C 12:22:56 Temperature Measurement, Probe Number = 3, T = 28.565000 C 12:22:56 Temperature Measurement, Probe Number = 4, T = 28.778000 C 12:22:57 Temperature Measurement, Probe Number = 5, T = 25.062000 C 12:22:57 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.37 P3 = RTD Outside, Value = 29.13 P4 = Supply Water, Value = 28.12 P5 = Return Water, Value = 28.57 P6 = @ RTD Outside, Value = 28.78 P7 = Ambient, Value = 25.06 12:22:57 3 Axis Hall Probe Field Measurement, Bx = 0.000010, By = 0.000020, Bz = -0.000050 T 12:23:04 Move Stage, position = -0.080000 12:23:06 Stage Position Measurement, pos = -0.0801050 m 12:23:07 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:23:08 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:23:08 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:23:09 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:23:09 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:23:10 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:23:10 Main Volt Voltage Measurement, V = 0.000002 V 12:23:11 Temperature Measurement, Probe Number = 0, T = 29.370000 C 12:23:15 Temperature Measurement, Probe Number = 1, T = 29.131000 C 12:23:15 Temperature Measurement, Probe Number = 2, T = 28.141000 C 12:23:16 Temperature Measurement, Probe Number = 3, T = 28.524000 C 12:23:16 Temperature Measurement, Probe Number = 4, T = 28.780000 C 12:23:17 Temperature Measurement, Probe Number = 5, T = 25.007000 C 12:23:17 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.37 P3 = RTD Outside, Value = 29.13 P4 = Supply Water, Value = 28.14 P5 = Return Water, Value = 28.52 P6 = @ RTD Outside, Value = 28.78 P7 = Ambient, Value = 25.01 12:23:17 3 Axis Hall Probe Field Measurement, Bx = 0.000010, By = 0.000010, Bz = -0.000060 T 12:23:25 Move Stage, position = -0.078000 12:23:26 Stage Position Measurement, pos = -0.0781250 m 12:23:27 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:23:28 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:23:28 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:23:29 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:23:29 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:23:30 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:23:30 Main Volt Voltage Measurement, V = 0.000002 V 12:23:31 Temperature Measurement, Probe Number = 0, T = 29.369000 C 12:23:35 Temperature Measurement, Probe Number = 1, T = 29.135000 C 12:23:35 Temperature Measurement, Probe Number = 2, T = 28.145000 C 12:23:36 Temperature Measurement, Probe Number = 3, T = 28.497000 C 12:23:36 Temperature Measurement, Probe Number = 4, T = 28.782000 C 12:23:37 Temperature Measurement, Probe Number = 5, T = 24.966000 C 12:23:37 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.37 P3 = RTD Outside, Value = 29.14 P4 = Supply Water, Value = 28.14 P5 = Return Water, Value = 28.50 P6 = @ RTD Outside, Value = 28.78 P7 = Ambient, Value = 24.97 12:23:37 3 Axis Hall Probe Field Measurement, Bx = 0.000020, By = 0.000000, Bz = -0.000070 T 12:23:45 Move Stage, position = -0.076000 12:23:46 Stage Position Measurement, pos = -0.0761400 m 12:23:47 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:23:48 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:23:49 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:23:49 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:23:49 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:23:50 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:23:51 Main Volt Voltage Measurement, V = 0.000002 V 12:23:51 Temperature Measurement, Probe Number = 0, T = 29.367000 C 12:23:55 Temperature Measurement, Probe Number = 1, T = 29.130000 C 12:23:55 Temperature Measurement, Probe Number = 2, T = 28.109000 C 12:23:56 Temperature Measurement, Probe Number = 3, T = 28.505000 C 12:23:57 Temperature Measurement, Probe Number = 4, T = 28.781000 C 12:23:57 Temperature Measurement, Probe Number = 5, T = 25.021000 C 12:23:57 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.37 P3 = RTD Outside, Value = 29.13 P4 = Supply Water, Value = 28.11 P5 = Return Water, Value = 28.50 P6 = @ RTD Outside, Value = 28.78 P7 = Ambient, Value = 25.02 12:23:57 3 Axis Hall Probe Field Measurement, Bx = 0.000020, By = 0.000020, Bz = -0.000080 T 12:24:05 Move Stage, position = -0.074000 12:24:07 Stage Position Measurement, pos = -0.0741600 m 12:24:07 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:24:08 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:24:09 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:24:09 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:24:09 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:24:10 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:24:11 Main Volt Voltage Measurement, V = 0.000002 V 12:24:11 Temperature Measurement, Probe Number = 0, T = 29.365000 C 12:24:15 Temperature Measurement, Probe Number = 1, T = 29.133000 C 12:24:15 Temperature Measurement, Probe Number = 2, T = 28.059000 C 12:24:16 Temperature Measurement, Probe Number = 3, T = 28.507000 C 12:24:17 Temperature Measurement, Probe Number = 4, T = 28.783000 C 12:24:17 Temperature Measurement, Probe Number = 5, T = 25.040000 C 12:24:17 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.36 P3 = RTD Outside, Value = 29.13 P4 = Supply Water, Value = 28.06 P5 = Return Water, Value = 28.51 P6 = @ RTD Outside, Value = 28.78 P7 = Ambient, Value = 25.04 12:24:17 3 Axis Hall Probe Field Measurement, Bx = 0.000020, By = 0.000020, Bz = -0.000090 T 12:24:25 Move Stage, position = -0.072000 12:24:27 Stage Position Measurement, pos = -0.0721250 m 12:24:27 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:24:28 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:24:29 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:24:30 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:24:30 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:24:30 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:24:31 Main Volt Voltage Measurement, V = 0.000002 V 12:24:31 Temperature Measurement, Probe Number = 0, T = 29.364000 C 12:24:35 Temperature Measurement, Probe Number = 1, T = 29.128000 C 12:24:36 Temperature Measurement, Probe Number = 2, T = 28.014000 C 12:24:36 Temperature Measurement, Probe Number = 3, T = 28.506000 C 12:24:37 Temperature Measurement, Probe Number = 4, T = 28.779000 C 12:24:37 Temperature Measurement, Probe Number = 5, T = 25.091000 C 12:24:37 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.36 P3 = RTD Outside, Value = 29.13 P4 = Supply Water, Value = 28.01 P5 = Return Water, Value = 28.51 P6 = @ RTD Outside, Value = 28.78 P7 = Ambient, Value = 25.09 12:24:37 3 Axis Hall Probe Field Measurement, Bx = 0.000010, By = 0.000010, Bz = -0.000090 T 12:24:45 Move Stage, position = -0.070000 12:24:47 Stage Position Measurement, pos = -0.0701350 m 12:24:48 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:24:48 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:24:49 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:24:50 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:24:50 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:24:50 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:24:51 Main Volt Voltage Measurement, V = 0.000002 V 12:24:52 Temperature Measurement, Probe Number = 0, T = 29.368000 C 12:24:55 Temperature Measurement, Probe Number = 1, T = 29.133000 C 12:24:56 Temperature Measurement, Probe Number = 2, T = 28.009000 C 12:24:56 Temperature Measurement, Probe Number = 3, T = 28.508000 C 12:24:57 Temperature Measurement, Probe Number = 4, T = 28.784000 C 12:24:58 Temperature Measurement, Probe Number = 5, T = 25.081000 C 12:24:58 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.37 P3 = RTD Outside, Value = 29.13 P4 = Supply Water, Value = 28.01 P5 = Return Water, Value = 28.51 P6 = @ RTD Outside, Value = 28.78 P7 = Ambient, Value = 25.08 12:24:58 3 Axis Hall Probe Field Measurement, Bx = 0.000010, By = 0.000020, Bz = -0.000100 T 12:25:05 Move Stage, position = -0.068000 12:25:07 Stage Position Measurement, pos = -0.0681300 m 12:25:08 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:25:08 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:25:09 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:25:10 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:25:10 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:25:11 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:25:11 Main Volt Voltage Measurement, V = 0.000002 V 12:25:12 Temperature Measurement, Probe Number = 0, T = 29.372000 C 12:25:15 Temperature Measurement, Probe Number = 1, T = 29.136000 C 12:25:16 Temperature Measurement, Probe Number = 2, T = 28.042000 C 12:25:16 Temperature Measurement, Probe Number = 3, T = 28.521000 C 12:25:17 Temperature Measurement, Probe Number = 4, T = 28.786000 C 12:25:18 Temperature Measurement, Probe Number = 5, T = 25.062000 C 12:25:18 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.37 P3 = RTD Outside, Value = 29.14 P4 = Supply Water, Value = 28.04 P5 = Return Water, Value = 28.52 P6 = @ RTD Outside, Value = 28.79 P7 = Ambient, Value = 25.06 12:25:18 3 Axis Hall Probe Field Measurement, Bx = 0.000010, By = 0.000010, Bz = -0.000110 T 12:25:25 Move Stage, position = -0.066000 12:25:27 Stage Position Measurement, pos = -0.0661250 m 12:25:28 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:25:28 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:25:29 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:25:30 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:25:30 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:25:31 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:25:31 Main Volt Voltage Measurement, V = 0.000002 V 12:25:32 Temperature Measurement, Probe Number = 0, T = 29.375000 C 12:25:35 Temperature Measurement, Probe Number = 1, T = 29.139000 C 12:25:36 Temperature Measurement, Probe Number = 2, T = 28.105000 C 12:25:37 Temperature Measurement, Probe Number = 3, T = 28.538000 C 12:25:37 Temperature Measurement, Probe Number = 4, T = 28.788000 C 12:25:38 Temperature Measurement, Probe Number = 5, T = 25.052000 C 12:25:38 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.38 P3 = RTD Outside, Value = 29.14 P4 = Supply Water, Value = 28.11 P5 = Return Water, Value = 28.54 P6 = @ RTD Outside, Value = 28.79 P7 = Ambient, Value = 25.05 12:25:38 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000000, Bz = -0.000110 T 12:25:45 Move Stage, position = -0.064000 12:25:47 Stage Position Measurement, pos = -0.0640950 m 12:25:48 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:25:49 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:25:49 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:25:50 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:25:50 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:25:51 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:25:51 Main Volt Voltage Measurement, V = 0.000002 V 12:25:52 Temperature Measurement, Probe Number = 0, T = 29.379000 C 12:25:56 Temperature Measurement, Probe Number = 1, T = 29.136000 C 12:25:56 Temperature Measurement, Probe Number = 2, T = 28.167000 C 12:25:57 Temperature Measurement, Probe Number = 3, T = 28.560000 C 12:25:57 Temperature Measurement, Probe Number = 4, T = 28.792000 C 12:25:58 Temperature Measurement, Probe Number = 5, T = 25.030000 C 12:25:58 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.38 P3 = RTD Outside, Value = 29.14 P4 = Supply Water, Value = 28.17 P5 = Return Water, Value = 28.56 P6 = @ RTD Outside, Value = 28.79 P7 = Ambient, Value = 25.03 12:25:58 3 Axis Hall Probe Field Measurement, Bx = 0.000010, By = 0.000010, Bz = -0.000120 T 12:26:06 Move Stage, position = -0.062000 12:26:07 Stage Position Measurement, pos = -0.0620850 m 12:26:08 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:26:09 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:26:09 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:26:10 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:26:10 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:26:11 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:26:11 Main Volt Voltage Measurement, V = 0.000002 V 12:26:12 Temperature Measurement, Probe Number = 0, T = 29.378000 C 12:26:16 Temperature Measurement, Probe Number = 1, T = 29.139000 C 12:26:16 Temperature Measurement, Probe Number = 2, T = 28.197000 C 12:26:17 Temperature Measurement, Probe Number = 3, T = 28.523000 C 12:26:17 Temperature Measurement, Probe Number = 4, T = 28.793000 C 12:26:18 Temperature Measurement, Probe Number = 5, T = 24.909000 C 12:26:18 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.38 P3 = RTD Outside, Value = 29.14 P4 = Supply Water, Value = 28.20 P5 = Return Water, Value = 28.52 P6 = @ RTD Outside, Value = 28.79 P7 = Ambient, Value = 24.91 12:26:18 3 Axis Hall Probe Field Measurement, Bx = 0.000010, By = 0.000000, Bz = -0.000100 T 12:26:26 Move Stage, position = -0.060000 12:26:27 Stage Position Measurement, pos = -0.0601150 m 12:26:28 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:26:29 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:26:29 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:26:30 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:26:30 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:26:31 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:26:31 Main Volt Voltage Measurement, V = 0.000002 V 12:26:32 Temperature Measurement, Probe Number = 0, T = 29.375000 C 12:26:36 Temperature Measurement, Probe Number = 1, T = 29.135000 C 12:26:36 Temperature Measurement, Probe Number = 2, T = 28.206000 C 12:26:37 Temperature Measurement, Probe Number = 3, T = 28.469000 C 12:26:37 Temperature Measurement, Probe Number = 4, T = 28.793000 C 12:26:38 Temperature Measurement, Probe Number = 5, T = 24.973000 C 12:26:38 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.38 P3 = RTD Outside, Value = 29.14 P4 = Supply Water, Value = 28.21 P5 = Return Water, Value = 28.47 P6 = @ RTD Outside, Value = 28.79 P7 = Ambient, Value = 24.97 12:26:38 3 Axis Hall Probe Field Measurement, Bx = 0.000020, By = 0.000000, Bz = -0.000110 T 12:26:46 Move Stage, position = -0.058000 12:26:47 Stage Position Measurement, pos = -0.0581150 m 12:26:48 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:26:49 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:26:50 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:26:50 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:26:50 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:26:51 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:26:52 Main Volt Voltage Measurement, V = 0.000002 V 12:26:52 Temperature Measurement, Probe Number = 0, T = 29.374000 C 12:26:56 Temperature Measurement, Probe Number = 1, T = 29.138000 C 12:26:56 Temperature Measurement, Probe Number = 2, T = 28.228000 C 12:26:57 Temperature Measurement, Probe Number = 3, T = 28.456000 C 12:26:58 Temperature Measurement, Probe Number = 4, T = 28.794000 C 12:26:58 Temperature Measurement, Probe Number = 5, T = 25.042000 C 12:26:58 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.37 P3 = RTD Outside, Value = 29.14 P4 = Supply Water, Value = 28.23 P5 = Return Water, Value = 28.46 P6 = @ RTD Outside, Value = 28.79 P7 = Ambient, Value = 25.04 12:26:58 3 Axis Hall Probe Field Measurement, Bx = 0.000010, By = 0.000010, Bz = -0.000090 T 12:27:06 Move Stage, position = -0.056000 12:27:08 Stage Position Measurement, pos = -0.0561500 m 12:27:08 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:27:09 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:27:10 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:27:10 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:27:10 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:27:11 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:27:12 Main Volt Voltage Measurement, V = 0.000002 V 12:27:12 Temperature Measurement, Probe Number = 0, T = 29.375000 C 12:27:16 Temperature Measurement, Probe Number = 1, T = 29.140000 C 12:27:17 Temperature Measurement, Probe Number = 2, T = 28.271000 C 12:27:17 Temperature Measurement, Probe Number = 3, T = 28.471000 C 12:27:18 Temperature Measurement, Probe Number = 4, T = 28.795000 C 12:27:18 Temperature Measurement, Probe Number = 5, T = 25.073000 C 12:27:18 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.38 P3 = RTD Outside, Value = 29.14 P4 = Supply Water, Value = 28.27 P5 = Return Water, Value = 28.47 P6 = @ RTD Outside, Value = 28.80 P7 = Ambient, Value = 25.07 12:27:18 3 Axis Hall Probe Field Measurement, Bx = 0.000010, By = 0.000010, Bz = -0.000090 T 12:27:26 Move Stage, position = -0.054000 12:27:28 Stage Position Measurement, pos = -0.0541300 m 12:27:28 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:27:29 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:27:30 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:27:31 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:27:31 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:27:31 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:27:32 Main Volt Voltage Measurement, V = 0.000002 V 12:27:32 Temperature Measurement, Probe Number = 0, T = 29.377000 C 12:27:36 Temperature Measurement, Probe Number = 1, T = 29.136000 C 12:27:37 Temperature Measurement, Probe Number = 2, T = 28.317000 C 12:27:37 Temperature Measurement, Probe Number = 3, T = 28.508000 C 12:27:38 Temperature Measurement, Probe Number = 4, T = 28.793000 C 12:27:38 Temperature Measurement, Probe Number = 5, T = 25.071000 C 12:27:38 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.38 P3 = RTD Outside, Value = 29.14 P4 = Supply Water, Value = 28.32 P5 = Return Water, Value = 28.51 P6 = @ RTD Outside, Value = 28.79 P7 = Ambient, Value = 25.07 12:27:38 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000000, Bz = -0.000080 T 12:27:46 Move Stage, position = -0.052000 12:27:48 Stage Position Measurement, pos = -0.0521250 m 12:27:48 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:27:49 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:27:50 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:27:51 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:27:51 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:27:51 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:27:52 Main Volt Voltage Measurement, V = 0.000002 V 12:27:53 Temperature Measurement, Probe Number = 0, T = 29.378000 C 12:27:56 Temperature Measurement, Probe Number = 1, T = 29.143000 C 12:27:57 Temperature Measurement, Probe Number = 2, T = 28.304000 C 12:27:57 Temperature Measurement, Probe Number = 3, T = 28.540000 C 12:27:58 Temperature Measurement, Probe Number = 4, T = 28.795000 C 12:27:58 Temperature Measurement, Probe Number = 5, T = 25.028000 C 12:27:58 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.38 P3 = RTD Outside, Value = 29.14 P4 = Supply Water, Value = 28.30 P5 = Return Water, Value = 28.54 P6 = @ RTD Outside, Value = 28.80 P7 = Ambient, Value = 25.03 12:27:59 3 Axis Hall Probe Field Measurement, Bx = 0.000020, By = -0.000010, Bz = -0.000070 T 12:28:06 Move Stage, position = -0.050000 12:28:08 Stage Position Measurement, pos = -0.0501200 m 12:28:09 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:28:09 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:28:10 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:28:11 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:28:11 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:28:11 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:28:12 Main Volt Voltage Measurement, V = 0.000002 V 12:28:13 Temperature Measurement, Probe Number = 0, T = 29.383000 C 12:28:16 Temperature Measurement, Probe Number = 1, T = 29.146000 C 12:28:17 Temperature Measurement, Probe Number = 2, T = 28.288000 C 12:28:17 Temperature Measurement, Probe Number = 3, T = 28.555000 C 12:28:18 Temperature Measurement, Probe Number = 4, T = 28.798000 C 12:28:19 Temperature Measurement, Probe Number = 5, T = 25.102000 C 12:28:19 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.38 P3 = RTD Outside, Value = 29.15 P4 = Supply Water, Value = 28.29 P5 = Return Water, Value = 28.55 P6 = @ RTD Outside, Value = 28.80 P7 = Ambient, Value = 25.10 12:28:19 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000000, Bz = -0.000050 T 12:28:26 Move Stage, position = -0.048000 12:28:28 Stage Position Measurement, pos = -0.0481200 m 12:28:29 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:28:29 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:28:30 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:28:31 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:28:31 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:28:32 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:28:32 Main Volt Voltage Measurement, V = 0.000002 V 12:28:33 Temperature Measurement, Probe Number = 0, T = 29.389000 C 12:28:36 Temperature Measurement, Probe Number = 1, T = 29.151000 C 12:28:37 Temperature Measurement, Probe Number = 2, T = 28.284000 C 12:28:38 Temperature Measurement, Probe Number = 3, T = 28.578000 C 12:28:38 Temperature Measurement, Probe Number = 4, T = 28.803000 C 12:28:39 Temperature Measurement, Probe Number = 5, T = 25.169000 C 12:28:39 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.39 P3 = RTD Outside, Value = 29.15 P4 = Supply Water, Value = 28.28 P5 = Return Water, Value = 28.58 P6 = @ RTD Outside, Value = 28.80 P7 = Ambient, Value = 25.17 12:28:39 3 Axis Hall Probe Field Measurement, Bx = 0.000010, By = 0.000000, Bz = -0.000040 T 12:28:46 Move Stage, position = -0.046000 12:28:48 Stage Position Measurement, pos = -0.0461050 m 12:28:49 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:28:50 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:28:50 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:28:51 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:28:51 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:28:52 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:28:52 Main Volt Voltage Measurement, V = 0.000002 V 12:28:53 Temperature Measurement, Probe Number = 0, T = 29.389000 C 12:28:56 Temperature Measurement, Probe Number = 1, T = 29.152000 C 12:28:57 Temperature Measurement, Probe Number = 2, T = 28.293000 C 12:28:58 Temperature Measurement, Probe Number = 3, T = 28.587000 C 12:28:58 Temperature Measurement, Probe Number = 4, T = 28.807000 C 12:28:59 Temperature Measurement, Probe Number = 5, T = 25.032000 C 12:28:59 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.39 P3 = RTD Outside, Value = 29.15 P4 = Supply Water, Value = 28.29 P5 = Return Water, Value = 28.59 P6 = @ RTD Outside, Value = 28.81 P7 = Ambient, Value = 25.03 12:28:59 3 Axis Hall Probe Field Measurement, Bx = 0.000010, By = 0.000000, Bz = -0.000030 T 12:29:07 Move Stage, position = -0.044000 12:29:08 Stage Position Measurement, pos = -0.0440800 m 12:29:09 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:29:10 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:29:10 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:29:11 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:29:11 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:29:12 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:29:12 Main Volt Voltage Measurement, V = 0.000002 V 12:29:13 Temperature Measurement, Probe Number = 0, T = 29.390000 C 12:29:17 Temperature Measurement, Probe Number = 1, T = 29.147000 C 12:29:17 Temperature Measurement, Probe Number = 2, T = 28.286000 C 12:29:18 Temperature Measurement, Probe Number = 3, T = 28.587000 C 12:29:18 Temperature Measurement, Probe Number = 4, T = 28.805000 C 12:29:19 Temperature Measurement, Probe Number = 5, T = 25.135000 C 12:29:19 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.39 P3 = RTD Outside, Value = 29.15 P4 = Supply Water, Value = 28.29 P5 = Return Water, Value = 28.59 P6 = @ RTD Outside, Value = 28.80 P7 = Ambient, Value = 25.14 12:29:19 3 Axis Hall Probe Field Measurement, Bx = 0.000010, By = 0.000010, Bz = -0.000010 T 12:29:27 Move Stage, position = -0.042000 12:29:28 Stage Position Measurement, pos = -0.0421000 m 12:29:29 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:29:30 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:29:30 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:29:31 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:29:31 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:29:32 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:29:32 Main Volt Voltage Measurement, V = 0.000002 V 12:29:33 Temperature Measurement, Probe Number = 0, T = 29.387000 C 12:29:37 Temperature Measurement, Probe Number = 1, T = 29.146000 C 12:29:37 Temperature Measurement, Probe Number = 2, T = 28.274000 C 12:29:38 Temperature Measurement, Probe Number = 3, T = 28.604000 C 12:29:38 Temperature Measurement, Probe Number = 4, T = 28.798000 C 12:29:39 Temperature Measurement, Probe Number = 5, T = 25.127000 C 12:29:39 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.39 P3 = RTD Outside, Value = 29.15 P4 = Supply Water, Value = 28.27 P5 = Return Water, Value = 28.60 P6 = @ RTD Outside, Value = 28.80 P7 = Ambient, Value = 25.13 12:29:39 3 Axis Hall Probe Field Measurement, Bx = 0.000010, By = 0.000010, Bz = 0.000000 T 12:29:47 Move Stage, position = -0.040000 12:29:48 Stage Position Measurement, pos = -0.0401050 m 12:29:49 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:29:50 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:29:51 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:29:51 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:29:51 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:29:52 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:29:53 Main Volt Voltage Measurement, V = 0.000002 V 12:29:53 Temperature Measurement, Probe Number = 0, T = 29.391000 C 12:29:57 Temperature Measurement, Probe Number = 1, T = 29.153000 C 12:29:57 Temperature Measurement, Probe Number = 2, T = 28.290000 C 12:29:58 Temperature Measurement, Probe Number = 3, T = 28.614000 C 12:29:59 Temperature Measurement, Probe Number = 4, T = 28.802000 C 12:29:59 Temperature Measurement, Probe Number = 5, T = 25.138000 C 12:29:59 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.39 P3 = RTD Outside, Value = 29.15 P4 = Supply Water, Value = 28.29 P5 = Return Water, Value = 28.61 P6 = @ RTD Outside, Value = 28.80 P7 = Ambient, Value = 25.14 12:29:59 3 Axis Hall Probe Field Measurement, Bx = 0.000010, By = 0.000010, Bz = 0.000000 T 12:30:07 Move Stage, position = -0.038000 12:30:09 Stage Position Measurement, pos = -0.0381300 m 12:30:09 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:30:10 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:30:11 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:30:11 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:30:11 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:30:12 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:30:13 Main Volt Voltage Measurement, V = 0.000002 V 12:30:13 Temperature Measurement, Probe Number = 0, T = 29.395000 C 12:30:17 Temperature Measurement, Probe Number = 1, T = 29.150000 C 12:30:17 Temperature Measurement, Probe Number = 2, T = 28.324000 C 12:30:18 Temperature Measurement, Probe Number = 3, T = 28.629000 C 12:30:19 Temperature Measurement, Probe Number = 4, T = 28.806000 C 12:30:19 Temperature Measurement, Probe Number = 5, T = 25.120000 C 12:30:19 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.39 P3 = RTD Outside, Value = 29.15 P4 = Supply Water, Value = 28.32 P5 = Return Water, Value = 28.63 P6 = @ RTD Outside, Value = 28.81 P7 = Ambient, Value = 25.12 12:30:19 3 Axis Hall Probe Field Measurement, Bx = 0.000010, By = 0.000010, Bz = 0.000000 T 12:30:27 Move Stage, position = -0.036000 12:30:29 Stage Position Measurement, pos = -0.0361250 m 12:30:29 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:30:30 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:30:31 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:30:31 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:30:31 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:30:32 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:30:33 Main Volt Voltage Measurement, V = 0.000002 V 12:30:33 Temperature Measurement, Probe Number = 0, T = 29.395000 C 12:30:37 Temperature Measurement, Probe Number = 1, T = 29.155000 C 12:30:38 Temperature Measurement, Probe Number = 2, T = 28.324000 C 12:30:38 Temperature Measurement, Probe Number = 3, T = 28.622000 C 12:30:39 Temperature Measurement, Probe Number = 4, T = 28.809000 C 12:30:39 Temperature Measurement, Probe Number = 5, T = 25.040000 C 12:30:39 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.39 P3 = RTD Outside, Value = 29.16 P4 = Supply Water, Value = 28.32 P5 = Return Water, Value = 28.62 P6 = @ RTD Outside, Value = 28.81 P7 = Ambient, Value = 25.04 12:30:39 3 Axis Hall Probe Field Measurement, Bx = 0.000010, By = 0.000000, Bz = 0.000010 T 12:30:47 Move Stage, position = -0.034000 12:30:49 Stage Position Measurement, pos = -0.0341100 m 12:30:50 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:30:50 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:30:51 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:30:52 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:30:52 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:30:52 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:30:53 Main Volt Voltage Measurement, V = 0.000002 V 12:30:54 Temperature Measurement, Probe Number = 0, T = 29.396000 C 12:30:57 Temperature Measurement, Probe Number = 1, T = 29.159000 C 12:30:58 Temperature Measurement, Probe Number = 2, T = 28.271000 C 12:30:58 Temperature Measurement, Probe Number = 3, T = 28.609000 C 12:30:59 Temperature Measurement, Probe Number = 4, T = 28.813000 C 12:31:00 Temperature Measurement, Probe Number = 5, T = 25.029000 C 12:31:00 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.40 P3 = RTD Outside, Value = 29.16 P4 = Supply Water, Value = 28.27 P5 = Return Water, Value = 28.61 P6 = @ RTD Outside, Value = 28.81 P7 = Ambient, Value = 25.03 12:31:00 3 Axis Hall Probe Field Measurement, Bx = 0.000010, By = 0.000010, Bz = 0.000010 T 12:31:07 Move Stage, position = -0.032000 12:31:09 Stage Position Measurement, pos = -0.0321000 m 12:31:10 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:31:10 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:31:11 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:31:12 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:31:12 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:31:12 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:31:13 Main Volt Voltage Measurement, V = 0.000002 V 12:31:14 Temperature Measurement, Probe Number = 0, T = 29.393000 C 12:31:17 Temperature Measurement, Probe Number = 1, T = 29.158000 C 12:31:18 Temperature Measurement, Probe Number = 2, T = 28.233000 C 12:31:18 Temperature Measurement, Probe Number = 3, T = 28.580000 C 12:31:19 Temperature Measurement, Probe Number = 4, T = 28.815000 C 12:31:20 Temperature Measurement, Probe Number = 5, T = 24.958000 C 12:31:20 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.39 P3 = RTD Outside, Value = 29.16 P4 = Supply Water, Value = 28.23 P5 = Return Water, Value = 28.58 P6 = @ RTD Outside, Value = 28.82 P7 = Ambient, Value = 24.96 12:31:20 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000000, Bz = 0.000010 T 12:31:27 Move Stage, position = -0.030000 12:31:29 Stage Position Measurement, pos = -0.0300850 m 12:31:30 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:31:30 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:31:31 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:31:32 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:31:32 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:31:33 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:31:33 Main Volt Voltage Measurement, V = 0.000002 V 12:31:34 Temperature Measurement, Probe Number = 0, T = 29.388000 C 12:31:37 Temperature Measurement, Probe Number = 1, T = 29.150000 C 12:31:38 Temperature Measurement, Probe Number = 2, T = 28.228000 C 12:31:39 Temperature Measurement, Probe Number = 3, T = 28.536000 C 12:31:39 Temperature Measurement, Probe Number = 4, T = 28.814000 C 12:31:40 Temperature Measurement, Probe Number = 5, T = 24.979000 C 12:31:40 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.39 P3 = RTD Outside, Value = 29.15 P4 = Supply Water, Value = 28.23 P5 = Return Water, Value = 28.54 P6 = @ RTD Outside, Value = 28.81 P7 = Ambient, Value = 24.98 12:31:40 3 Axis Hall Probe Field Measurement, Bx = 0.000010, By = 0.000010, Bz = 0.000000 T 12:31:47 Move Stage, position = -0.028000 12:31:49 Stage Position Measurement, pos = -0.0280650 m 12:31:50 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:31:51 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:31:51 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:31:52 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:31:52 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:31:53 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:31:53 Main Volt Voltage Measurement, V = 0.000002 V 12:31:54 Temperature Measurement, Probe Number = 0, T = 29.386000 C 12:31:57 Temperature Measurement, Probe Number = 1, T = 29.147000 C 12:31:58 Temperature Measurement, Probe Number = 2, T = 28.242000 C 12:31:59 Temperature Measurement, Probe Number = 3, T = 28.518000 C 12:31:59 Temperature Measurement, Probe Number = 4, T = 28.813000 C 12:32:00 Temperature Measurement, Probe Number = 5, T = 24.963000 C 12:32:00 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.39 P3 = RTD Outside, Value = 29.15 P4 = Supply Water, Value = 28.24 P5 = Return Water, Value = 28.52 P6 = @ RTD Outside, Value = 28.81 P7 = Ambient, Value = 24.96 12:32:00 3 Axis Hall Probe Field Measurement, Bx = 0.000010, By = 0.000020, Bz = 0.000020 T 12:32:08 Move Stage, position = -0.026000 12:32:09 Stage Position Measurement, pos = -0.0260750 m 12:32:10 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:32:11 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:32:11 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:32:12 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:32:12 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:32:13 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:32:13 Main Volt Voltage Measurement, V = 0.000002 V 12:32:14 Temperature Measurement, Probe Number = 0, T = 29.384000 C 12:32:18 Temperature Measurement, Probe Number = 1, T = 29.151000 C 12:32:18 Temperature Measurement, Probe Number = 2, T = 28.263000 C 12:32:19 Temperature Measurement, Probe Number = 3, T = 28.497000 C 12:32:19 Temperature Measurement, Probe Number = 4, T = 28.814000 C 12:32:20 Temperature Measurement, Probe Number = 5, T = 24.960000 C 12:32:20 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.38 P3 = RTD Outside, Value = 29.15 P4 = Supply Water, Value = 28.26 P5 = Return Water, Value = 28.50 P6 = @ RTD Outside, Value = 28.81 P7 = Ambient, Value = 24.96 12:32:20 3 Axis Hall Probe Field Measurement, Bx = 0.000010, By = 0.000010, Bz = 0.000020 T 12:32:28 Move Stage, position = -0.024000 12:32:29 Stage Position Measurement, pos = -0.0240850 m 12:32:30 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:32:31 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:32:31 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:32:32 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:32:32 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:32:33 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:32:33 Main Volt Voltage Measurement, V = 0.000002 V 12:32:34 Temperature Measurement, Probe Number = 0, T = 29.385000 C 12:32:38 Temperature Measurement, Probe Number = 1, T = 29.147000 C 12:32:38 Temperature Measurement, Probe Number = 2, T = 28.289000 C 12:32:39 Temperature Measurement, Probe Number = 3, T = 28.493000 C 12:32:39 Temperature Measurement, Probe Number = 4, T = 28.813000 C 12:32:40 Temperature Measurement, Probe Number = 5, T = 25.054000 C 12:32:40 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.39 P3 = RTD Outside, Value = 29.15 P4 = Supply Water, Value = 28.29 P5 = Return Water, Value = 28.49 P6 = @ RTD Outside, Value = 28.81 P7 = Ambient, Value = 25.05 12:32:40 3 Axis Hall Probe Field Measurement, Bx = 0.000010, By = 0.000010, Bz = 0.000010 T 12:32:48 Move Stage, position = -0.022000 12:32:49 Stage Position Measurement, pos = -0.0220700 m 12:32:50 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:32:51 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:32:52 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:32:52 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:32:52 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:32:53 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:32:54 Main Volt Voltage Measurement, V = 0.000002 V 12:32:54 Temperature Measurement, Probe Number = 0, T = 29.384000 C 12:32:58 Temperature Measurement, Probe Number = 1, T = 29.143000 C 12:32:58 Temperature Measurement, Probe Number = 2, T = 28.299000 C 12:32:59 Temperature Measurement, Probe Number = 3, T = 28.515000 C 12:33:00 Temperature Measurement, Probe Number = 4, T = 28.800000 C 12:33:00 Temperature Measurement, Probe Number = 5, T = 25.183000 C 12:33:00 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.38 P3 = RTD Outside, Value = 29.14 P4 = Supply Water, Value = 28.30 P5 = Return Water, Value = 28.52 P6 = @ RTD Outside, Value = 28.80 P7 = Ambient, Value = 25.18 12:33:00 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000010, Bz = 0.000010 T 12:33:08 Move Stage, position = -0.020000 12:33:10 Stage Position Measurement, pos = -0.0200950 m 12:33:10 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:33:11 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:33:12 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:33:12 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:33:12 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:33:13 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:33:14 Main Volt Voltage Measurement, V = 0.000002 V 12:33:14 Temperature Measurement, Probe Number = 0, T = 29.387000 C 12:33:18 Temperature Measurement, Probe Number = 1, T = 29.148000 C 12:33:19 Temperature Measurement, Probe Number = 2, T = 28.275000 C 12:33:19 Temperature Measurement, Probe Number = 3, T = 28.545000 C 12:33:20 Temperature Measurement, Probe Number = 4, T = 28.789000 C 12:33:20 Temperature Measurement, Probe Number = 5, T = 25.177000 C 12:33:20 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.39 P3 = RTD Outside, Value = 29.15 P4 = Supply Water, Value = 28.27 P5 = Return Water, Value = 28.55 P6 = @ RTD Outside, Value = 28.79 P7 = Ambient, Value = 25.18 12:33:20 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000000, Bz = 0.000010 T 12:33:28 Move Stage, position = -0.018000 12:33:30 Stage Position Measurement, pos = -0.0180900 m 12:33:30 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:33:31 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:33:32 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:33:33 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:33:33 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:33:33 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:33:34 Main Volt Voltage Measurement, V = 0.000002 V 12:33:34 Temperature Measurement, Probe Number = 0, T = 29.393000 C 12:33:38 Temperature Measurement, Probe Number = 1, T = 29.152000 C 12:33:39 Temperature Measurement, Probe Number = 2, T = 28.275000 C 12:33:39 Temperature Measurement, Probe Number = 3, T = 28.563000 C 12:33:40 Temperature Measurement, Probe Number = 4, T = 28.792000 C 12:33:40 Temperature Measurement, Probe Number = 5, T = 25.303000 C 12:33:40 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.39 P3 = RTD Outside, Value = 29.15 P4 = Supply Water, Value = 28.27 P5 = Return Water, Value = 28.56 P6 = @ RTD Outside, Value = 28.79 P7 = Ambient, Value = 25.30 12:33:41 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000000, Bz = 0.000010 T 12:33:48 Move Stage, position = -0.016000 12:33:50 Stage Position Measurement, pos = -0.0160900 m 12:33:51 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:33:51 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:33:52 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:33:53 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:33:53 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:33:53 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:33:54 Main Volt Voltage Measurement, V = 0.000002 V 12:33:55 Temperature Measurement, Probe Number = 0, T = 29.397000 C 12:33:58 Temperature Measurement, Probe Number = 1, T = 29.151000 C 12:33:59 Temperature Measurement, Probe Number = 2, T = 28.308000 C 12:33:59 Temperature Measurement, Probe Number = 3, T = 28.582000 C 12:34:00 Temperature Measurement, Probe Number = 4, T = 28.792000 C 12:34:01 Temperature Measurement, Probe Number = 5, T = 25.302000 C 12:34:01 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.40 P3 = RTD Outside, Value = 29.15 P4 = Supply Water, Value = 28.31 P5 = Return Water, Value = 28.58 P6 = @ RTD Outside, Value = 28.79 P7 = Ambient, Value = 25.30 12:34:01 3 Axis Hall Probe Field Measurement, Bx = -0.000010, By = 0.000000, Bz = 0.000010 T 12:34:08 Move Stage, position = -0.014000 12:34:10 Stage Position Measurement, pos = -0.0140650 m 12:34:11 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:34:11 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:34:12 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:34:13 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:34:13 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:34:14 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:34:14 Main Volt Voltage Measurement, V = 0.000002 V 12:34:15 Temperature Measurement, Probe Number = 0, T = 29.401000 C 12:34:18 Temperature Measurement, Probe Number = 1, T = 29.153000 C 12:34:19 Temperature Measurement, Probe Number = 2, T = 28.332000 C 12:34:19 Temperature Measurement, Probe Number = 3, T = 28.602000 C 12:34:20 Temperature Measurement, Probe Number = 4, T = 28.792000 C 12:34:21 Temperature Measurement, Probe Number = 5, T = 25.448000 C 12:34:21 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.40 P3 = RTD Outside, Value = 29.15 P4 = Supply Water, Value = 28.33 P5 = Return Water, Value = 28.60 P6 = @ RTD Outside, Value = 28.79 P7 = Ambient, Value = 25.45 12:34:21 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000010, Bz = 0.000010 T 12:34:28 Move Stage, position = -0.012000 12:34:30 Stage Position Measurement, pos = -0.0120600 m 12:34:31 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:34:31 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:34:32 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:34:33 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:34:33 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:34:34 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:34:34 Main Volt Voltage Measurement, V = 0.000002 V 12:34:35 Temperature Measurement, Probe Number = 0, T = 29.406000 C 12:34:38 Temperature Measurement, Probe Number = 1, T = 29.153000 C 12:34:39 Temperature Measurement, Probe Number = 2, T = 28.352000 C 12:34:40 Temperature Measurement, Probe Number = 3, T = 28.634000 C 12:34:40 Temperature Measurement, Probe Number = 4, T = 28.786000 C 12:34:41 Temperature Measurement, Probe Number = 5, T = 25.541000 C 12:34:41 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.41 P3 = RTD Outside, Value = 29.15 P4 = Supply Water, Value = 28.35 P5 = Return Water, Value = 28.63 P6 = @ RTD Outside, Value = 28.79 P7 = Ambient, Value = 25.54 12:34:41 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000010, Bz = 0.000010 T 12:34:49 Move Stage, position = -0.010000 12:34:50 Stage Position Measurement, pos = -0.0100350 m 12:34:51 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:34:52 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:34:52 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:34:53 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:34:53 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:34:54 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000059 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:34:54 Main Volt Voltage Measurement, V = 0.000002 V 12:34:55 Temperature Measurement, Probe Number = 0, T = 29.410000 C 12:34:59 Temperature Measurement, Probe Number = 1, T = 29.152000 C 12:34:59 Temperature Measurement, Probe Number = 2, T = 28.338000 C 12:35:00 Temperature Measurement, Probe Number = 3, T = 28.664000 C 12:35:00 Temperature Measurement, Probe Number = 4, T = 28.776000 C 12:35:01 Temperature Measurement, Probe Number = 5, T = 25.535000 C 12:35:01 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.41 P3 = RTD Outside, Value = 29.15 P4 = Supply Water, Value = 28.34 P5 = Return Water, Value = 28.66 P6 = @ RTD Outside, Value = 28.78 P7 = Ambient, Value = 25.54 12:35:01 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000000, Bz = 0.000030 T 12:35:09 Move Stage, position = -0.008000 12:35:10 Stage Position Measurement, pos = -0.0080400 m 12:35:11 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:35:12 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:35:12 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:35:13 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:35:13 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:35:14 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:35:14 Main Volt Voltage Measurement, V = 0.000002 V 12:35:15 Temperature Measurement, Probe Number = 0, T = 29.414000 C 12:35:19 Temperature Measurement, Probe Number = 1, T = 29.165000 C 12:35:19 Temperature Measurement, Probe Number = 2, T = 28.312000 C 12:35:20 Temperature Measurement, Probe Number = 3, T = 28.667000 C 12:35:20 Temperature Measurement, Probe Number = 4, T = 28.774000 C 12:35:21 Temperature Measurement, Probe Number = 5, T = 25.476000 C 12:35:21 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.41 P3 = RTD Outside, Value = 29.16 P4 = Supply Water, Value = 28.31 P5 = Return Water, Value = 28.67 P6 = @ RTD Outside, Value = 28.77 P7 = Ambient, Value = 25.48 12:35:21 3 Axis Hall Probe Field Measurement, Bx = 0.000010, By = 0.000010, Bz = 0.000030 T 12:35:29 Move Stage, position = -0.006000 12:35:30 Stage Position Measurement, pos = -0.0060550 m 12:35:31 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:35:32 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:35:33 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:35:33 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:35:33 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:35:34 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:35:34 Main Volt Voltage Measurement, V = 0.000002 V 12:35:35 Temperature Measurement, Probe Number = 0, T = 29.416000 C 12:35:39 Temperature Measurement, Probe Number = 1, T = 29.173000 C 12:35:39 Temperature Measurement, Probe Number = 2, T = 28.301000 C 12:35:40 Temperature Measurement, Probe Number = 3, T = 28.661000 C 12:35:40 Temperature Measurement, Probe Number = 4, T = 28.788000 C 12:35:41 Temperature Measurement, Probe Number = 5, T = 25.490000 C 12:35:41 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.42 P3 = RTD Outside, Value = 29.17 P4 = Supply Water, Value = 28.30 P5 = Return Water, Value = 28.66 P6 = @ RTD Outside, Value = 28.79 P7 = Ambient, Value = 25.49 12:35:41 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000000, Bz = 0.000020 T 12:35:49 Move Stage, position = -0.004000 12:35:50 Stage Position Measurement, pos = -0.0040400 m 12:35:51 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:35:52 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:35:53 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:35:53 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:35:53 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:35:54 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:35:55 Main Volt Voltage Measurement, V = 0.000002 V 12:35:55 Temperature Measurement, Probe Number = 0, T = 29.413000 C 12:35:59 Temperature Measurement, Probe Number = 1, T = 29.172000 C 12:35:59 Temperature Measurement, Probe Number = 2, T = 28.325000 C 12:36:00 Temperature Measurement, Probe Number = 3, T = 28.666000 C 12:36:01 Temperature Measurement, Probe Number = 4, T = 28.797000 C 12:36:01 Temperature Measurement, Probe Number = 5, T = 25.564000 C 12:36:01 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.41 P3 = RTD Outside, Value = 29.17 P4 = Supply Water, Value = 28.32 P5 = Return Water, Value = 28.67 P6 = @ RTD Outside, Value = 28.80 P7 = Ambient, Value = 25.56 12:36:01 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000000, Bz = 0.000030 T 12:36:09 Move Stage, position = -0.002000 12:36:11 Stage Position Measurement, pos = -0.0020500 m 12:36:11 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:36:12 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:36:13 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:36:13 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:36:13 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:36:14 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:36:15 Main Volt Voltage Measurement, V = 0.000002 V 12:36:15 Temperature Measurement, Probe Number = 0, T = 29.412000 C 12:36:19 Temperature Measurement, Probe Number = 1, T = 29.174000 C 12:36:20 Temperature Measurement, Probe Number = 2, T = 28.321000 C 12:36:20 Temperature Measurement, Probe Number = 3, T = 28.677000 C 12:36:21 Temperature Measurement, Probe Number = 4, T = 28.797000 C 12:36:21 Temperature Measurement, Probe Number = 5, T = 25.518000 C 12:36:21 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.41 P3 = RTD Outside, Value = 29.17 P4 = Supply Water, Value = 28.32 P5 = Return Water, Value = 28.68 P6 = @ RTD Outside, Value = 28.80 P7 = Ambient, Value = 25.52 12:36:21 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000010, Bz = 0.000030 T 12:36:29 Move Stage, position = 0.000000 12:36:31 Stage Position Measurement, pos = -0.0000350 m 12:36:31 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:36:32 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:36:33 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:36:34 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:36:34 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:36:34 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:36:35 Main Volt Voltage Measurement, V = 0.000002 V 12:36:35 Temperature Measurement, Probe Number = 0, T = 29.415000 C 12:36:39 Temperature Measurement, Probe Number = 1, T = 29.177000 C 12:36:40 Temperature Measurement, Probe Number = 2, T = 28.327000 C 12:36:40 Temperature Measurement, Probe Number = 3, T = 28.678000 C 12:36:41 Temperature Measurement, Probe Number = 4, T = 28.807000 C 12:36:41 Temperature Measurement, Probe Number = 5, T = 25.487000 C 12:36:41 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.41 P3 = RTD Outside, Value = 29.18 P4 = Supply Water, Value = 28.33 P5 = Return Water, Value = 28.68 P6 = @ RTD Outside, Value = 28.81 P7 = Ambient, Value = 25.49 12:36:41 3 Axis Hall Probe Field Measurement, Bx = -0.000010, By = 0.000010, Bz = 0.000020 T 12:36:49 Move Stage, position = 0.002000 12:36:51 Stage Position Measurement, pos = 0.0019600 m 12:36:52 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:36:52 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:36:53 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:36:54 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:36:54 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:36:54 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:36:55 Main Volt Voltage Measurement, V = 0.000002 V 12:36:56 Temperature Measurement, Probe Number = 0, T = 29.415000 C 12:36:59 Temperature Measurement, Probe Number = 1, T = 29.179000 C 12:37:00 Temperature Measurement, Probe Number = 2, T = 28.347000 C 12:37:00 Temperature Measurement, Probe Number = 3, T = 28.685000 C 12:37:01 Temperature Measurement, Probe Number = 4, T = 28.816000 C 12:37:02 Temperature Measurement, Probe Number = 5, T = 25.530000 C 12:37:02 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.41 P3 = RTD Outside, Value = 29.18 P4 = Supply Water, Value = 28.35 P5 = Return Water, Value = 28.68 P6 = @ RTD Outside, Value = 28.82 P7 = Ambient, Value = 25.53 12:37:02 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000010, Bz = 0.000020 T 12:37:09 Move Stage, position = 0.004000 12:37:11 Stage Position Measurement, pos = 0.0039800 m 12:37:12 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:37:12 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:37:13 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:37:14 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:37:14 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:37:15 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:37:15 Main Volt Voltage Measurement, V = 0.000002 V 12:37:16 Temperature Measurement, Probe Number = 0, T = 29.415000 C 12:37:19 Temperature Measurement, Probe Number = 1, T = 29.181000 C 12:37:20 Temperature Measurement, Probe Number = 2, T = 28.358000 C 12:37:20 Temperature Measurement, Probe Number = 3, T = 28.691000 C 12:37:21 Temperature Measurement, Probe Number = 4, T = 28.823000 C 12:37:22 Temperature Measurement, Probe Number = 5, T = 25.566000 C 12:37:22 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.41 P3 = RTD Outside, Value = 29.18 P4 = Supply Water, Value = 28.36 P5 = Return Water, Value = 28.69 P6 = @ RTD Outside, Value = 28.82 P7 = Ambient, Value = 25.57 12:37:22 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000010, Bz = 0.000020 T 12:37:29 Move Stage, position = 0.006000 12:37:31 Stage Position Measurement, pos = 0.0060100 m 12:37:32 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:37:32 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:37:33 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:37:34 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:37:34 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:37:35 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:37:35 Main Volt Voltage Measurement, V = 0.000002 V 12:37:36 Temperature Measurement, Probe Number = 0, T = 29.417000 C 12:37:39 Temperature Measurement, Probe Number = 1, T = 29.177000 C 12:37:40 Temperature Measurement, Probe Number = 2, T = 28.336000 C 12:37:41 Temperature Measurement, Probe Number = 3, T = 28.687000 C 12:37:41 Temperature Measurement, Probe Number = 4, T = 28.830000 C 12:37:42 Temperature Measurement, Probe Number = 5, T = 25.563000 C 12:37:42 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.42 P3 = RTD Outside, Value = 29.18 P4 = Supply Water, Value = 28.34 P5 = Return Water, Value = 28.69 P6 = @ RTD Outside, Value = 28.83 P7 = Ambient, Value = 25.56 12:37:42 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000000, Bz = 0.000030 T 12:37:50 Move Stage, position = 0.008000 12:37:51 Stage Position Measurement, pos = 0.0080300 m 12:37:52 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:37:53 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:37:53 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:37:54 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:37:54 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:37:55 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:37:55 Main Volt Voltage Measurement, V = 0.000002 V 12:37:56 Temperature Measurement, Probe Number = 0, T = 29.414000 C 12:38:00 Temperature Measurement, Probe Number = 1, T = 29.178000 C 12:38:00 Temperature Measurement, Probe Number = 2, T = 28.311000 C 12:38:01 Temperature Measurement, Probe Number = 3, T = 28.684000 C 12:38:01 Temperature Measurement, Probe Number = 4, T = 28.834000 C 12:38:02 Temperature Measurement, Probe Number = 5, T = 25.525000 C 12:38:02 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.41 P3 = RTD Outside, Value = 29.18 P4 = Supply Water, Value = 28.31 P5 = Return Water, Value = 28.68 P6 = @ RTD Outside, Value = 28.83 P7 = Ambient, Value = 25.52 12:38:02 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000010, Bz = 0.000020 T 12:38:10 Move Stage, position = 0.010000 12:38:11 Stage Position Measurement, pos = 0.0100350 m 12:38:12 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:38:13 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:38:13 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:38:14 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:38:14 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:38:15 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:38:15 Main Volt Voltage Measurement, V = 0.000002 V 12:38:16 Temperature Measurement, Probe Number = 0, T = 29.417000 C 12:38:20 Temperature Measurement, Probe Number = 1, T = 29.179000 C 12:38:20 Temperature Measurement, Probe Number = 2, T = 28.326000 C 12:38:21 Temperature Measurement, Probe Number = 3, T = 28.699000 C 12:38:21 Temperature Measurement, Probe Number = 4, T = 28.836000 C 12:38:22 Temperature Measurement, Probe Number = 5, T = 25.618000 C 12:38:22 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.42 P3 = RTD Outside, Value = 29.18 P4 = Supply Water, Value = 28.33 P5 = Return Water, Value = 28.70 P6 = @ RTD Outside, Value = 28.84 P7 = Ambient, Value = 25.62 12:38:22 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000010, Bz = 0.000030 T 12:38:30 Move Stage, position = 0.012000 12:38:31 Stage Position Measurement, pos = 0.0120100 m 12:38:32 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:38:33 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:38:34 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:38:34 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:38:34 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:38:35 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:38:36 Main Volt Voltage Measurement, V = 0.000002 V 12:38:36 Temperature Measurement, Probe Number = 0, T = 29.416000 C 12:38:40 Temperature Measurement, Probe Number = 1, T = 29.178000 C 12:38:40 Temperature Measurement, Probe Number = 2, T = 28.343000 C 12:38:41 Temperature Measurement, Probe Number = 3, T = 28.710000 C 12:38:41 Temperature Measurement, Probe Number = 4, T = 28.829000 C 12:38:42 Temperature Measurement, Probe Number = 5, T = 25.600000 C 12:38:42 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.42 P3 = RTD Outside, Value = 29.18 P4 = Supply Water, Value = 28.34 P5 = Return Water, Value = 28.71 P6 = @ RTD Outside, Value = 28.83 P7 = Ambient, Value = 25.60 12:38:42 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000010, Bz = 0.000030 T 12:38:50 Move Stage, position = 0.014000 12:38:51 Stage Position Measurement, pos = 0.0140150 m 12:38:52 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:38:53 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:38:54 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:38:54 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:38:54 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:38:55 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:38:56 Main Volt Voltage Measurement, V = 0.000002 V 12:38:56 Temperature Measurement, Probe Number = 0, T = 29.414000 C 12:39:00 Temperature Measurement, Probe Number = 1, T = 29.178000 C 12:39:00 Temperature Measurement, Probe Number = 2, T = 28.339000 C 12:39:01 Temperature Measurement, Probe Number = 3, T = 28.713000 C 12:39:02 Temperature Measurement, Probe Number = 4, T = 28.829000 C 12:39:02 Temperature Measurement, Probe Number = 5, T = 25.579000 C 12:39:02 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.41 P3 = RTD Outside, Value = 29.18 P4 = Supply Water, Value = 28.34 P5 = Return Water, Value = 28.71 P6 = @ RTD Outside, Value = 28.83 P7 = Ambient, Value = 25.58 12:39:02 3 Axis Hall Probe Field Measurement, Bx = -0.000010, By = 0.000010, Bz = 0.000030 T 12:39:10 Move Stage, position = 0.016000 12:39:12 Stage Position Measurement, pos = 0.0160150 m 12:39:12 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:39:13 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:39:14 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:39:14 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:39:14 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:39:15 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:39:16 Main Volt Voltage Measurement, V = 0.000002 V 12:39:16 Temperature Measurement, Probe Number = 0, T = 29.418000 C 12:39:20 Temperature Measurement, Probe Number = 1, T = 29.179000 C 12:39:21 Temperature Measurement, Probe Number = 2, T = 28.372000 C 12:39:21 Temperature Measurement, Probe Number = 3, T = 28.721000 C 12:39:22 Temperature Measurement, Probe Number = 4, T = 28.833000 C 12:39:22 Temperature Measurement, Probe Number = 5, T = 25.731000 C 12:39:22 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.42 P3 = RTD Outside, Value = 29.18 P4 = Supply Water, Value = 28.37 P5 = Return Water, Value = 28.72 P6 = @ RTD Outside, Value = 28.83 P7 = Ambient, Value = 25.73 12:39:22 3 Axis Hall Probe Field Measurement, Bx = -0.000020, By = 0.000000, Bz = 0.000020 T 12:39:30 Move Stage, position = 0.018000 12:39:32 Stage Position Measurement, pos = 0.0180200 m 12:39:32 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:39:33 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:39:34 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:39:35 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:39:35 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:39:35 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:39:36 Main Volt Voltage Measurement, V = 0.000002 V 12:39:36 Temperature Measurement, Probe Number = 0, T = 29.420000 C 12:39:40 Temperature Measurement, Probe Number = 1, T = 29.179000 C 12:39:41 Temperature Measurement, Probe Number = 2, T = 28.410000 C 12:39:41 Temperature Measurement, Probe Number = 3, T = 28.735000 C 12:39:42 Temperature Measurement, Probe Number = 4, T = 28.835000 C 12:39:42 Temperature Measurement, Probe Number = 5, T = 25.642000 C 12:39:42 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.42 P3 = RTD Outside, Value = 29.18 P4 = Supply Water, Value = 28.41 P5 = Return Water, Value = 28.73 P6 = @ RTD Outside, Value = 28.84 P7 = Ambient, Value = 25.64 12:39:42 3 Axis Hall Probe Field Measurement, Bx = -0.000010, By = 0.000010, Bz = 0.000010 T 12:39:50 Move Stage, position = 0.020000 12:39:52 Stage Position Measurement, pos = 0.0200250 m 12:39:53 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:39:53 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:39:54 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:39:55 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:39:55 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:39:55 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:39:56 Main Volt Voltage Measurement, V = 0.000002 V 12:39:57 Temperature Measurement, Probe Number = 0, T = 29.424000 C 12:40:00 Temperature Measurement, Probe Number = 1, T = 29.184000 C 12:40:01 Temperature Measurement, Probe Number = 2, T = 28.422000 C 12:40:01 Temperature Measurement, Probe Number = 3, T = 28.747000 C 12:40:02 Temperature Measurement, Probe Number = 4, T = 28.840000 C 12:40:03 Temperature Measurement, Probe Number = 5, T = 25.628000 C 12:40:03 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.42 P3 = RTD Outside, Value = 29.18 P4 = Supply Water, Value = 28.42 P5 = Return Water, Value = 28.75 P6 = @ RTD Outside, Value = 28.84 P7 = Ambient, Value = 25.63 12:40:03 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000010, Bz = 0.000020 T 12:40:10 Move Stage, position = 0.022000 12:40:12 Stage Position Measurement, pos = 0.0220400 m 12:40:13 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:40:13 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:40:14 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:40:15 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:40:15 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:40:16 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:40:16 Main Volt Voltage Measurement, V = 0.000002 V 12:40:17 Temperature Measurement, Probe Number = 0, T = 29.427000 C 12:40:20 Temperature Measurement, Probe Number = 1, T = 29.188000 C 12:40:21 Temperature Measurement, Probe Number = 2, T = 28.436000 C 12:40:21 Temperature Measurement, Probe Number = 3, T = 28.759000 C 12:40:22 Temperature Measurement, Probe Number = 4, T = 28.847000 C 12:40:23 Temperature Measurement, Probe Number = 5, T = 25.631000 C 12:40:23 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.43 P3 = RTD Outside, Value = 29.19 P4 = Supply Water, Value = 28.44 P5 = Return Water, Value = 28.76 P6 = @ RTD Outside, Value = 28.85 P7 = Ambient, Value = 25.63 12:40:23 3 Axis Hall Probe Field Measurement, Bx = -0.000010, By = 0.000010, Bz = 0.000020 T 12:40:30 Move Stage, position = 0.024000 12:40:32 Stage Position Measurement, pos = 0.0240300 m 12:40:33 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:40:33 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:40:34 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:40:35 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:40:35 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:40:36 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:40:36 Main Volt Voltage Measurement, V = 0.000002 V 12:40:37 Temperature Measurement, Probe Number = 0, T = 29.426000 C 12:40:41 Temperature Measurement, Probe Number = 1, T = 29.188000 C 12:40:41 Temperature Measurement, Probe Number = 2, T = 28.462000 C 12:40:42 Temperature Measurement, Probe Number = 3, T = 28.765000 C 12:40:42 Temperature Measurement, Probe Number = 4, T = 28.849000 C 12:40:43 Temperature Measurement, Probe Number = 5, T = 25.732000 C 12:40:43 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.43 P3 = RTD Outside, Value = 29.19 P4 = Supply Water, Value = 28.46 P5 = Return Water, Value = 28.77 P6 = @ RTD Outside, Value = 28.85 P7 = Ambient, Value = 25.73 12:40:43 3 Axis Hall Probe Field Measurement, Bx = -0.000010, By = 0.000010, Bz = 0.000010 T 12:40:51 Move Stage, position = 0.026000 12:40:52 Stage Position Measurement, pos = 0.0260700 m 12:40:53 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:40:54 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:40:54 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:40:55 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:40:55 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:40:56 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:40:56 Main Volt Voltage Measurement, V = 0.000001 V 12:40:57 Temperature Measurement, Probe Number = 0, T = 29.429000 C 12:41:01 Temperature Measurement, Probe Number = 1, T = 29.185000 C 12:41:01 Temperature Measurement, Probe Number = 2, T = 28.477000 C 12:41:02 Temperature Measurement, Probe Number = 3, T = 28.767000 C 12:41:02 Temperature Measurement, Probe Number = 4, T = 28.849000 C 12:41:03 Temperature Measurement, Probe Number = 5, T = 25.690000 C 12:41:03 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.43 P3 = RTD Outside, Value = 29.18 P4 = Supply Water, Value = 28.48 P5 = Return Water, Value = 28.77 P6 = @ RTD Outside, Value = 28.85 P7 = Ambient, Value = 25.69 12:41:03 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000010, Bz = 0.000020 T 12:41:11 Move Stage, position = 0.028000 12:41:12 Stage Position Measurement, pos = 0.0280750 m 12:41:13 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:41:14 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:41:15 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:41:15 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:41:15 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:41:16 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:41:17 Main Volt Voltage Measurement, V = 0.000001 V 12:41:17 Temperature Measurement, Probe Number = 0, T = 29.432000 C 12:41:21 Temperature Measurement, Probe Number = 1, T = 29.185000 C 12:41:21 Temperature Measurement, Probe Number = 2, T = 28.451000 C 12:41:22 Temperature Measurement, Probe Number = 3, T = 28.769000 C 12:41:23 Temperature Measurement, Probe Number = 4, T = 28.844000 C 12:41:23 Temperature Measurement, Probe Number = 5, T = 25.702000 C 12:41:23 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.43 P3 = RTD Outside, Value = 29.18 P4 = Supply Water, Value = 28.45 P5 = Return Water, Value = 28.77 P6 = @ RTD Outside, Value = 28.84 P7 = Ambient, Value = 25.70 12:41:23 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000020, Bz = 0.000010 T 12:41:31 Move Stage, position = 0.030000 12:41:33 Stage Position Measurement, pos = 0.0300550 m 12:41:33 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:41:34 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:41:35 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:41:35 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:41:35 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:41:36 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:41:37 Main Volt Voltage Measurement, V = 0.000002 V 12:41:37 Temperature Measurement, Probe Number = 0, T = 29.434000 C 12:41:41 Temperature Measurement, Probe Number = 1, T = 29.192000 C 12:41:42 Temperature Measurement, Probe Number = 2, T = 28.452000 C 12:41:42 Temperature Measurement, Probe Number = 3, T = 28.772000 C 12:41:43 Temperature Measurement, Probe Number = 4, T = 28.846000 C 12:41:43 Temperature Measurement, Probe Number = 5, T = 25.735000 C 12:41:43 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.43 P3 = RTD Outside, Value = 29.19 P4 = Supply Water, Value = 28.45 P5 = Return Water, Value = 28.77 P6 = @ RTD Outside, Value = 28.85 P7 = Ambient, Value = 25.73 12:41:43 3 Axis Hall Probe Field Measurement, Bx = -0.000010, By = 0.000010, Bz = 0.000010 T 12:41:51 Move Stage, position = 0.032000 12:41:53 Stage Position Measurement, pos = 0.0320550 m 12:41:53 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:41:54 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:41:55 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:41:56 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:41:56 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:41:56 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:41:57 Main Volt Voltage Measurement, V = 0.000002 V 12:41:57 Temperature Measurement, Probe Number = 0, T = 29.434000 C 12:42:01 Temperature Measurement, Probe Number = 1, T = 29.198000 C 12:42:02 Temperature Measurement, Probe Number = 2, T = 28.449000 C 12:42:02 Temperature Measurement, Probe Number = 3, T = 28.768000 C 12:42:03 Temperature Measurement, Probe Number = 4, T = 28.852000 C 12:42:03 Temperature Measurement, Probe Number = 5, T = 25.713000 C 12:42:03 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.43 P3 = RTD Outside, Value = 29.20 P4 = Supply Water, Value = 28.45 P5 = Return Water, Value = 28.77 P6 = @ RTD Outside, Value = 28.85 P7 = Ambient, Value = 25.71 12:42:03 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000010, Bz = 0.000020 T 12:42:11 Move Stage, position = 0.034000 12:42:13 Stage Position Measurement, pos = 0.0340400 m 12:42:13 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:42:14 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:42:15 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:42:16 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:42:16 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:42:16 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:42:17 Main Volt Voltage Measurement, V = 0.000002 V 12:42:18 Temperature Measurement, Probe Number = 0, T = 29.437000 C 12:42:21 Temperature Measurement, Probe Number = 1, T = 29.199000 C 12:42:22 Temperature Measurement, Probe Number = 2, T = 28.465000 C 12:42:22 Temperature Measurement, Probe Number = 3, T = 28.752000 C 12:42:23 Temperature Measurement, Probe Number = 4, T = 28.858000 C 12:42:23 Temperature Measurement, Probe Number = 5, T = 25.741000 C 12:42:23 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.44 P3 = RTD Outside, Value = 29.20 P4 = Supply Water, Value = 28.46 P5 = Return Water, Value = 28.75 P6 = @ RTD Outside, Value = 28.86 P7 = Ambient, Value = 25.74 12:42:24 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000000, Bz = 0.000010 T 12:42:31 Move Stage, position = 0.036000 12:42:33 Stage Position Measurement, pos = 0.0360550 m 12:42:34 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:42:34 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:42:35 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:42:36 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:42:36 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:42:36 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:42:37 Main Volt Voltage Measurement, V = 0.000002 V 12:42:38 Temperature Measurement, Probe Number = 0, T = 29.441000 C 12:42:41 Temperature Measurement, Probe Number = 1, T = 29.202000 C 12:42:42 Temperature Measurement, Probe Number = 2, T = 28.484000 C 12:42:42 Temperature Measurement, Probe Number = 3, T = 28.750000 C 12:42:43 Temperature Measurement, Probe Number = 4, T = 28.865000 C 12:42:44 Temperature Measurement, Probe Number = 5, T = 25.750000 C 12:42:44 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.44 P3 = RTD Outside, Value = 29.20 P4 = Supply Water, Value = 28.48 P5 = Return Water, Value = 28.75 P6 = @ RTD Outside, Value = 28.86 P7 = Ambient, Value = 25.75 12:42:44 3 Axis Hall Probe Field Measurement, Bx = -0.000010, By = 0.000000, Bz = 0.000020 T 12:42:51 Move Stage, position = 0.038000 12:42:53 Stage Position Measurement, pos = 0.0380650 m 12:42:54 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:42:54 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:42:55 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:42:56 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:42:56 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:42:57 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:42:57 Main Volt Voltage Measurement, V = 0.000002 V 12:42:58 Temperature Measurement, Probe Number = 0, T = 29.446000 C 12:43:01 Temperature Measurement, Probe Number = 1, T = 29.206000 C 12:43:02 Temperature Measurement, Probe Number = 2, T = 28.491000 C 12:43:03 Temperature Measurement, Probe Number = 3, T = 28.760000 C 12:43:03 Temperature Measurement, Probe Number = 4, T = 28.869000 C 12:43:04 Temperature Measurement, Probe Number = 5, T = 25.731000 C 12:43:04 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.45 P3 = RTD Outside, Value = 29.21 P4 = Supply Water, Value = 28.49 P5 = Return Water, Value = 28.76 P6 = @ RTD Outside, Value = 28.87 P7 = Ambient, Value = 25.73 12:43:04 3 Axis Hall Probe Field Measurement, Bx = -0.000010, By = 0.000010, Bz = 0.000020 T 12:43:11 Move Stage, position = 0.040000 12:43:13 Stage Position Measurement, pos = 0.0400800 m 12:43:14 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:43:15 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:43:15 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:43:16 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:43:16 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:43:17 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:43:17 Main Volt Voltage Measurement, V = 0.000001 V 12:43:18 Temperature Measurement, Probe Number = 0, T = 29.448000 C 12:43:21 Temperature Measurement, Probe Number = 1, T = 29.208000 C 12:43:22 Temperature Measurement, Probe Number = 2, T = 28.496000 C 12:43:23 Temperature Measurement, Probe Number = 3, T = 28.771000 C 12:43:23 Temperature Measurement, Probe Number = 4, T = 28.874000 C 12:43:24 Temperature Measurement, Probe Number = 5, T = 25.730000 C 12:43:24 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.45 P3 = RTD Outside, Value = 29.21 P4 = Supply Water, Value = 28.50 P5 = Return Water, Value = 28.77 P6 = @ RTD Outside, Value = 28.87 P7 = Ambient, Value = 25.73 12:43:24 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000010, Bz = 0.000010 T 12:43:32 Move Stage, position = 0.042000 12:43:33 Stage Position Measurement, pos = 0.0420750 m 12:43:34 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:43:35 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:43:35 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:43:36 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:43:36 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:43:37 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:43:37 Main Volt Voltage Measurement, V = 0.000001 V 12:43:38 Temperature Measurement, Probe Number = 0, T = 29.449000 C 12:43:42 Temperature Measurement, Probe Number = 1, T = 29.211000 C 12:43:42 Temperature Measurement, Probe Number = 2, T = 28.484000 C 12:43:43 Temperature Measurement, Probe Number = 3, T = 28.779000 C 12:43:43 Temperature Measurement, Probe Number = 4, T = 28.880000 C 12:43:44 Temperature Measurement, Probe Number = 5, T = 25.663000 C 12:43:44 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.45 P3 = RTD Outside, Value = 29.21 P4 = Supply Water, Value = 28.48 P5 = Return Water, Value = 28.78 P6 = @ RTD Outside, Value = 28.88 P7 = Ambient, Value = 25.66 12:43:44 3 Axis Hall Probe Field Measurement, Bx = -0.000010, By = 0.000010, Bz = 0.000010 T 12:43:52 Move Stage, position = 0.044000 12:43:53 Stage Position Measurement, pos = 0.0440950 m 12:43:54 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:43:55 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:43:56 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:43:56 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:43:56 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:43:57 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:43:58 Main Volt Voltage Measurement, V = 0.000001 V 12:43:58 Temperature Measurement, Probe Number = 0, T = 29.448000 C 12:44:02 Temperature Measurement, Probe Number = 1, T = 29.213000 C 12:44:02 Temperature Measurement, Probe Number = 2, T = 28.440000 C 12:44:03 Temperature Measurement, Probe Number = 3, T = 28.755000 C 12:44:04 Temperature Measurement, Probe Number = 4, T = 28.888000 C 12:44:04 Temperature Measurement, Probe Number = 5, T = 25.669000 C 12:44:04 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.45 P3 = RTD Outside, Value = 29.21 P4 = Supply Water, Value = 28.44 P5 = Return Water, Value = 28.75 P6 = @ RTD Outside, Value = 28.89 P7 = Ambient, Value = 25.67 12:44:04 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000010, Bz = 0.000010 T 12:44:12 Move Stage, position = 0.046000 12:44:13 Stage Position Measurement, pos = 0.0461000 m 12:44:14 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:44:15 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:44:16 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:44:16 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:44:16 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:44:17 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:44:18 Main Volt Voltage Measurement, V = 0.000001 V 12:44:18 Temperature Measurement, Probe Number = 0, T = 29.447000 C 12:44:22 Temperature Measurement, Probe Number = 1, T = 29.217000 C 12:44:22 Temperature Measurement, Probe Number = 2, T = 28.399000 C 12:44:23 Temperature Measurement, Probe Number = 3, T = 28.740000 C 12:44:24 Temperature Measurement, Probe Number = 4, T = 28.894000 C 12:44:24 Temperature Measurement, Probe Number = 5, T = 25.654000 C 12:44:24 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.45 P3 = RTD Outside, Value = 29.22 P4 = Supply Water, Value = 28.40 P5 = Return Water, Value = 28.74 P6 = @ RTD Outside, Value = 28.89 P7 = Ambient, Value = 25.65 12:44:24 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000020, Bz = 0.000010 T 12:44:32 Move Stage, position = 0.048000 12:44:34 Stage Position Measurement, pos = 0.0480900 m 12:44:34 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:44:35 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:44:36 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:44:36 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:44:36 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:44:37 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:44:38 Main Volt Voltage Measurement, V = 0.000001 V 12:44:38 Temperature Measurement, Probe Number = 0, T = 29.447000 C 12:44:42 Temperature Measurement, Probe Number = 1, T = 29.218000 C 12:44:43 Temperature Measurement, Probe Number = 2, T = 28.402000 C 12:44:43 Temperature Measurement, Probe Number = 3, T = 28.738000 C 12:44:44 Temperature Measurement, Probe Number = 4, T = 28.896000 C 12:44:44 Temperature Measurement, Probe Number = 5, T = 25.599000 C 12:44:44 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.45 P3 = RTD Outside, Value = 29.22 P4 = Supply Water, Value = 28.40 P5 = Return Water, Value = 28.74 P6 = @ RTD Outside, Value = 28.90 P7 = Ambient, Value = 25.60 12:44:44 3 Axis Hall Probe Field Measurement, Bx = -0.000010, By = 0.000020, Bz = 0.000010 T 12:44:52 Move Stage, position = 0.050000 12:44:54 Stage Position Measurement, pos = 0.0500850 m 12:44:54 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:44:55 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:44:56 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:44:57 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:44:57 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:44:57 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:44:58 Main Volt Voltage Measurement, V = 0.000002 V 12:44:58 Temperature Measurement, Probe Number = 0, T = 29.447000 C 12:45:02 Temperature Measurement, Probe Number = 1, T = 29.212000 C 12:45:03 Temperature Measurement, Probe Number = 2, T = 28.416000 C 12:45:03 Temperature Measurement, Probe Number = 3, T = 28.710000 C 12:45:04 Temperature Measurement, Probe Number = 4, T = 28.895000 C 12:45:04 Temperature Measurement, Probe Number = 5, T = 25.520000 C 12:45:04 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.45 P3 = RTD Outside, Value = 29.21 P4 = Supply Water, Value = 28.42 P5 = Return Water, Value = 28.71 P6 = @ RTD Outside, Value = 28.89 P7 = Ambient, Value = 25.52 12:45:04 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000010, Bz = 0.000010 T 12:45:12 Move Stage, position = 0.052000 12:45:14 Stage Position Measurement, pos = 0.0520650 m 12:45:15 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:45:15 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:45:16 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:45:17 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:45:17 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:45:17 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:45:18 Main Volt Voltage Measurement, V = 0.000002 V 12:45:19 Temperature Measurement, Probe Number = 0, T = 29.448000 C 12:45:22 Temperature Measurement, Probe Number = 1, T = 29.207000 C 12:45:23 Temperature Measurement, Probe Number = 2, T = 28.448000 C 12:45:23 Temperature Measurement, Probe Number = 3, T = 28.706000 C 12:45:24 Temperature Measurement, Probe Number = 4, T = 28.892000 C 12:45:25 Temperature Measurement, Probe Number = 5, T = 25.625000 C 12:45:25 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.45 P3 = RTD Outside, Value = 29.21 P4 = Supply Water, Value = 28.45 P5 = Return Water, Value = 28.71 P6 = @ RTD Outside, Value = 28.89 P7 = Ambient, Value = 25.62 12:45:25 3 Axis Hall Probe Field Measurement, Bx = -0.000010, By = 0.000010, Bz = 0.000000 T 12:45:32 Move Stage, position = 0.054000 12:45:34 Stage Position Measurement, pos = 0.0540600 m 12:45:35 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:45:35 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:45:36 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:45:37 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:45:37 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:45:38 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:45:38 Main Volt Voltage Measurement, V = 0.000002 V 12:45:39 Temperature Measurement, Probe Number = 0, T = 29.445000 C 12:45:42 Temperature Measurement, Probe Number = 1, T = 29.206000 C 12:45:43 Temperature Measurement, Probe Number = 2, T = 28.493000 C 12:45:43 Temperature Measurement, Probe Number = 3, T = 28.711000 C 12:45:44 Temperature Measurement, Probe Number = 4, T = 28.889000 C 12:45:45 Temperature Measurement, Probe Number = 5, T = 25.665000 C 12:45:45 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.45 P3 = RTD Outside, Value = 29.21 P4 = Supply Water, Value = 28.49 P5 = Return Water, Value = 28.71 P6 = @ RTD Outside, Value = 28.89 P7 = Ambient, Value = 25.66 12:45:45 3 Axis Hall Probe Field Measurement, Bx = -0.000010, By = 0.000010, Bz = 0.000000 T 12:45:52 Move Stage, position = 0.056000 12:45:54 Stage Position Measurement, pos = 0.0560600 m 12:45:55 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:45:55 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:45:56 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:45:57 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:45:57 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:45:58 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:45:58 Main Volt Voltage Measurement, V = 0.000002 V 12:45:59 Temperature Measurement, Probe Number = 0, T = 29.446000 C 12:46:02 Temperature Measurement, Probe Number = 1, T = 29.212000 C 12:46:03 Temperature Measurement, Probe Number = 2, T = 28.521000 C 12:46:04 Temperature Measurement, Probe Number = 3, T = 28.730000 C 12:46:04 Temperature Measurement, Probe Number = 4, T = 28.892000 C 12:46:05 Temperature Measurement, Probe Number = 5, T = 25.648000 C 12:46:05 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.45 P3 = RTD Outside, Value = 29.21 P4 = Supply Water, Value = 28.52 P5 = Return Water, Value = 28.73 P6 = @ RTD Outside, Value = 28.89 P7 = Ambient, Value = 25.65 12:46:05 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000010, Bz = 0.000010 T 12:46:13 Move Stage, position = 0.058000 12:46:14 Stage Position Measurement, pos = 0.0580850 m 12:46:15 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:46:16 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:46:16 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:46:17 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:46:17 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:46:18 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:46:18 Main Volt Voltage Measurement, V = 0.000002 V 12:46:19 Temperature Measurement, Probe Number = 0, T = 29.446000 C 12:46:23 Temperature Measurement, Probe Number = 1, T = 29.214000 C 12:46:23 Temperature Measurement, Probe Number = 2, T = 28.530000 C 12:46:24 Temperature Measurement, Probe Number = 3, T = 28.753000 C 12:46:24 Temperature Measurement, Probe Number = 4, T = 28.898000 C 12:46:25 Temperature Measurement, Probe Number = 5, T = 25.594000 C 12:46:25 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.45 P3 = RTD Outside, Value = 29.21 P4 = Supply Water, Value = 28.53 P5 = Return Water, Value = 28.75 P6 = @ RTD Outside, Value = 28.90 P7 = Ambient, Value = 25.59 12:46:25 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000020, Bz = 0.000000 T 12:46:33 Move Stage, position = 0.060000 12:46:34 Stage Position Measurement, pos = 0.0600900 m 12:46:35 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:46:36 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:46:36 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:46:37 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:46:37 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:46:38 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:46:38 Main Volt Voltage Measurement, V = 0.000002 V 12:46:39 Temperature Measurement, Probe Number = 0, T = 29.444000 C 12:46:43 Temperature Measurement, Probe Number = 1, T = 29.213000 C 12:46:43 Temperature Measurement, Probe Number = 2, T = 28.509000 C 12:46:44 Temperature Measurement, Probe Number = 3, T = 28.739000 C 12:46:44 Temperature Measurement, Probe Number = 4, T = 28.900000 C 12:46:45 Temperature Measurement, Probe Number = 5, T = 25.554000 C 12:46:45 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.44 P3 = RTD Outside, Value = 29.21 P4 = Supply Water, Value = 28.51 P5 = Return Water, Value = 28.74 P6 = @ RTD Outside, Value = 28.90 P7 = Ambient, Value = 25.55 12:46:45 3 Axis Hall Probe Field Measurement, Bx = 0.000010, By = 0.000010, Bz = 0.000000 T 12:46:53 Move Stage, position = 0.062000 12:46:54 Stage Position Measurement, pos = 0.0621150 m 12:46:55 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:46:56 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:46:57 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:46:57 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:46:57 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:46:58 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:46:59 Main Volt Voltage Measurement, V = 0.000001 V 12:46:59 Temperature Measurement, Probe Number = 0, T = 29.441000 C 12:47:03 Temperature Measurement, Probe Number = 1, T = 29.210000 C 12:47:03 Temperature Measurement, Probe Number = 2, T = 28.489000 C 12:47:04 Temperature Measurement, Probe Number = 3, T = 28.726000 C 12:47:04 Temperature Measurement, Probe Number = 4, T = 28.900000 C 12:47:05 Temperature Measurement, Probe Number = 5, T = 25.723000 C 12:47:05 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.44 P3 = RTD Outside, Value = 29.21 P4 = Supply Water, Value = 28.49 P5 = Return Water, Value = 28.73 P6 = @ RTD Outside, Value = 28.90 P7 = Ambient, Value = 25.72 12:47:05 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000010, Bz = -0.000010 T 12:47:13 Move Stage, position = 0.064000 12:47:14 Stage Position Measurement, pos = 0.0640950 m 12:47:15 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:47:16 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:47:17 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:47:17 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:47:17 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:47:18 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:47:19 Main Volt Voltage Measurement, V = 0.000002 V 12:47:19 Temperature Measurement, Probe Number = 0, T = 29.440000 C 12:47:23 Temperature Measurement, Probe Number = 1, T = 29.204000 C 12:47:23 Temperature Measurement, Probe Number = 2, T = 28.504000 C 12:47:24 Temperature Measurement, Probe Number = 3, T = 28.717000 C 12:47:25 Temperature Measurement, Probe Number = 4, T = 28.896000 C 12:47:25 Temperature Measurement, Probe Number = 5, T = 25.722000 C 12:47:25 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.44 P3 = RTD Outside, Value = 29.20 P4 = Supply Water, Value = 28.50 P5 = Return Water, Value = 28.72 P6 = @ RTD Outside, Value = 28.90 P7 = Ambient, Value = 25.72 12:47:25 3 Axis Hall Probe Field Measurement, Bx = -0.000010, By = 0.000010, Bz = -0.000020 T 12:47:33 Move Stage, position = 0.066000 12:47:35 Stage Position Measurement, pos = 0.0660950 m 12:47:35 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:47:36 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:47:37 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:47:37 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:47:37 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:47:38 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:47:39 Main Volt Voltage Measurement, V = 0.000001 V 12:47:39 Temperature Measurement, Probe Number = 0, T = 29.440000 C 12:47:43 Temperature Measurement, Probe Number = 1, T = 29.203000 C 12:47:44 Temperature Measurement, Probe Number = 2, T = 28.538000 C 12:47:44 Temperature Measurement, Probe Number = 3, T = 28.719000 C 12:47:45 Temperature Measurement, Probe Number = 4, T = 28.891000 C 12:47:45 Temperature Measurement, Probe Number = 5, T = 25.696000 C 12:47:45 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.44 P3 = RTD Outside, Value = 29.20 P4 = Supply Water, Value = 28.54 P5 = Return Water, Value = 28.72 P6 = @ RTD Outside, Value = 28.89 P7 = Ambient, Value = 25.70 12:47:45 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000020, Bz = -0.000010 T 12:47:53 Move Stage, position = 0.068000 12:47:55 Stage Position Measurement, pos = 0.0680900 m 12:47:55 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:47:56 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:47:57 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:47:58 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:47:58 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:47:58 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:47:59 Main Volt Voltage Measurement, V = 0.000002 V 12:48:00 Temperature Measurement, Probe Number = 0, T = 29.440000 C 12:48:03 Temperature Measurement, Probe Number = 1, T = 29.202000 C 12:48:04 Temperature Measurement, Probe Number = 2, T = 28.544000 C 12:48:04 Temperature Measurement, Probe Number = 3, T = 28.739000 C 12:48:05 Temperature Measurement, Probe Number = 4, T = 28.887000 C 12:48:06 Temperature Measurement, Probe Number = 5, T = 25.677000 C 12:48:06 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.44 P3 = RTD Outside, Value = 29.20 P4 = Supply Water, Value = 28.54 P5 = Return Water, Value = 28.74 P6 = @ RTD Outside, Value = 28.89 P7 = Ambient, Value = 25.68 12:48:06 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000020, Bz = -0.000020 T 12:48:13 Move Stage, position = 0.070000 12:48:15 Stage Position Measurement, pos = 0.0700750 m 12:48:16 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:48:16 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:48:17 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:48:18 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:48:18 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:48:18 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:48:19 Main Volt Voltage Measurement, V = 0.000001 V 12:48:20 Temperature Measurement, Probe Number = 0, T = 29.441000 C 12:48:23 Temperature Measurement, Probe Number = 1, T = 29.207000 C 12:48:24 Temperature Measurement, Probe Number = 2, T = 28.520000 C 12:48:24 Temperature Measurement, Probe Number = 3, T = 28.750000 C 12:48:25 Temperature Measurement, Probe Number = 4, T = 28.886000 C 12:48:26 Temperature Measurement, Probe Number = 5, T = 25.683000 C 12:48:26 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.44 P3 = RTD Outside, Value = 29.21 P4 = Supply Water, Value = 28.52 P5 = Return Water, Value = 28.75 P6 = @ RTD Outside, Value = 28.89 P7 = Ambient, Value = 25.68 12:48:26 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000020, Bz = -0.000030 T 12:48:33 Move Stage, position = 0.072000 12:48:35 Stage Position Measurement, pos = 0.0720800 m 12:48:36 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:48:36 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:48:37 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:48:38 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:48:38 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:48:39 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:48:39 Main Volt Voltage Measurement, V = 0.000001 V 12:48:40 Temperature Measurement, Probe Number = 0, T = 29.446000 C 12:48:43 Temperature Measurement, Probe Number = 1, T = 29.210000 C 12:48:44 Temperature Measurement, Probe Number = 2, T = 28.494000 C 12:48:45 Temperature Measurement, Probe Number = 3, T = 28.750000 C 12:48:45 Temperature Measurement, Probe Number = 4, T = 28.890000 C 12:48:46 Temperature Measurement, Probe Number = 5, T = 25.759000 C 12:48:46 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.45 P3 = RTD Outside, Value = 29.21 P4 = Supply Water, Value = 28.49 P5 = Return Water, Value = 28.75 P6 = @ RTD Outside, Value = 28.89 P7 = Ambient, Value = 25.76 12:48:46 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000020, Bz = -0.000030 T 12:48:53 Move Stage, position = 0.074000 12:48:55 Stage Position Measurement, pos = 0.0740700 m 12:48:56 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:48:57 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:48:57 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:48:58 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:48:58 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:48:59 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:48:59 Main Volt Voltage Measurement, V = 0.000002 V 12:49:00 Temperature Measurement, Probe Number = 0, T = 29.448000 C 12:49:03 Temperature Measurement, Probe Number = 1, T = 29.211000 C 12:49:04 Temperature Measurement, Probe Number = 2, T = 28.486000 C 12:49:05 Temperature Measurement, Probe Number = 3, T = 28.751000 C 12:49:05 Temperature Measurement, Probe Number = 4, T = 28.895000 C 12:49:06 Temperature Measurement, Probe Number = 5, T = 25.747000 C 12:49:06 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.45 P3 = RTD Outside, Value = 29.21 P4 = Supply Water, Value = 28.49 P5 = Return Water, Value = 28.75 P6 = @ RTD Outside, Value = 28.89 P7 = Ambient, Value = 25.75 12:49:06 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000010, Bz = -0.000030 T 12:49:14 Move Stage, position = 0.076000 12:49:15 Stage Position Measurement, pos = 0.0761050 m 12:49:16 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:49:17 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:49:17 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:49:18 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:49:18 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:49:19 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:49:19 Main Volt Voltage Measurement, V = 0.000001 V 12:49:20 Temperature Measurement, Probe Number = 0, T = 29.450000 C 12:49:24 Temperature Measurement, Probe Number = 1, T = 29.211000 C 12:49:24 Temperature Measurement, Probe Number = 2, T = 28.513000 C 12:49:25 Temperature Measurement, Probe Number = 3, T = 28.762000 C 12:49:25 Temperature Measurement, Probe Number = 4, T = 28.897000 C 12:49:26 Temperature Measurement, Probe Number = 5, T = 25.780000 C 12:49:26 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.45 P3 = RTD Outside, Value = 29.21 P4 = Supply Water, Value = 28.51 P5 = Return Water, Value = 28.76 P6 = @ RTD Outside, Value = 28.90 P7 = Ambient, Value = 25.78 12:49:26 3 Axis Hall Probe Field Measurement, Bx = 0.000010, By = 0.000020, Bz = -0.000030 T 12:49:34 Move Stage, position = 0.078000 12:49:35 Stage Position Measurement, pos = 0.0781000 m 12:49:36 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:49:37 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:49:37 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:49:38 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:49:38 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:49:39 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:49:39 Main Volt Voltage Measurement, V = 0.000002 V 12:49:40 Temperature Measurement, Probe Number = 0, T = 29.451000 C 12:49:44 Temperature Measurement, Probe Number = 1, T = 29.217000 C 12:49:44 Temperature Measurement, Probe Number = 2, T = 28.509000 C 12:49:45 Temperature Measurement, Probe Number = 3, T = 28.768000 C 12:49:45 Temperature Measurement, Probe Number = 4, T = 28.901000 C 12:49:46 Temperature Measurement, Probe Number = 5, T = 25.725000 C 12:49:46 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.45 P3 = RTD Outside, Value = 29.22 P4 = Supply Water, Value = 28.51 P5 = Return Water, Value = 28.77 P6 = @ RTD Outside, Value = 28.90 P7 = Ambient, Value = 25.73 12:49:46 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000030, Bz = -0.000050 T 12:49:54 Move Stage, position = 0.080000 12:49:55 Stage Position Measurement, pos = 0.0801200 m 12:49:56 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:49:57 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:49:58 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:49:58 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:49:58 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:49:59 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:50:00 Main Volt Voltage Measurement, V = 0.000002 V 12:50:00 Temperature Measurement, Probe Number = 0, T = 29.449000 C 12:50:04 Temperature Measurement, Probe Number = 1, T = 29.217000 C 12:50:04 Temperature Measurement, Probe Number = 2, T = 28.477000 C 12:50:05 Temperature Measurement, Probe Number = 3, T = 28.769000 C 12:50:06 Temperature Measurement, Probe Number = 4, T = 28.906000 C 12:50:06 Temperature Measurement, Probe Number = 5, T = 25.678000 C 12:50:06 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.45 P3 = RTD Outside, Value = 29.22 P4 = Supply Water, Value = 28.48 P5 = Return Water, Value = 28.77 P6 = @ RTD Outside, Value = 28.91 P7 = Ambient, Value = 25.68 12:50:06 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000020, Bz = -0.000060 T 12:50:14 Move Stage, position = 0.082000 12:50:15 Stage Position Measurement, pos = 0.0821100 m 12:50:16 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:50:17 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:50:18 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:50:19 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:50:19 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:50:19 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:50:20 Main Volt Voltage Measurement, V = 0.000001 V 12:50:20 Temperature Measurement, Probe Number = 0, T = 29.446000 C 12:50:24 Temperature Measurement, Probe Number = 1, T = 29.210000 C 12:50:25 Temperature Measurement, Probe Number = 2, T = 28.447000 C 12:50:25 Temperature Measurement, Probe Number = 3, T = 28.766000 C 12:50:26 Temperature Measurement, Probe Number = 4, T = 28.899000 C 12:50:26 Temperature Measurement, Probe Number = 5, T = 25.707000 C 12:50:26 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.45 P3 = RTD Outside, Value = 29.21 P4 = Supply Water, Value = 28.45 P5 = Return Water, Value = 28.77 P6 = @ RTD Outside, Value = 28.90 P7 = Ambient, Value = 25.71 12:50:26 3 Axis Hall Probe Field Measurement, Bx = -0.000010, By = 0.000020, Bz = -0.000060 T 12:50:34 Move Stage, position = 0.084000 12:50:36 Stage Position Measurement, pos = 0.0840900 m 12:50:36 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:50:37 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:50:38 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:50:39 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:50:39 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:50:39 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:50:40 Main Volt Voltage Measurement, V = 0.000002 V 12:50:41 Temperature Measurement, Probe Number = 0, T = 29.446000 C 12:50:44 Temperature Measurement, Probe Number = 1, T = 29.214000 C 12:50:45 Temperature Measurement, Probe Number = 2, T = 28.419000 C 12:50:45 Temperature Measurement, Probe Number = 3, T = 28.764000 C 12:50:46 Temperature Measurement, Probe Number = 4, T = 28.896000 C 12:50:46 Temperature Measurement, Probe Number = 5, T = 25.705000 C 12:50:46 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.45 P3 = RTD Outside, Value = 29.21 P4 = Supply Water, Value = 28.42 P5 = Return Water, Value = 28.76 P6 = @ RTD Outside, Value = 28.90 P7 = Ambient, Value = 25.70 12:50:47 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000020, Bz = -0.000070 T 12:50:54 Move Stage, position = 0.086000 12:50:56 Stage Position Measurement, pos = 0.0860950 m 12:50:57 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:50:57 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:50:58 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:50:59 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:50:59 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:50:59 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:51:00 Main Volt Voltage Measurement, V = 0.000001 V 12:51:01 Temperature Measurement, Probe Number = 0, T = 29.443000 C 12:51:04 Temperature Measurement, Probe Number = 1, T = 29.201000 C 12:51:05 Temperature Measurement, Probe Number = 2, T = 28.385000 C 12:51:06 Temperature Measurement, Probe Number = 3, T = 28.768000 C 12:51:06 Temperature Measurement, Probe Number = 4, T = 28.889000 C 12:51:07 Temperature Measurement, Probe Number = 5, T = 25.557000 C 12:51:07 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.44 P3 = RTD Outside, Value = 29.20 P4 = Supply Water, Value = 28.39 P5 = Return Water, Value = 28.77 P6 = @ RTD Outside, Value = 28.89 P7 = Ambient, Value = 25.56 12:51:07 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000020, Bz = -0.000070 T 12:51:14 Move Stage, position = 0.088000 12:51:16 Stage Position Measurement, pos = 0.0880800 m 12:51:17 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:51:18 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:51:18 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:51:19 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:51:19 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:51:20 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:51:20 Main Volt Voltage Measurement, V = 0.000002 V 12:51:21 Temperature Measurement, Probe Number = 0, T = 29.444000 C 12:51:24 Temperature Measurement, Probe Number = 1, T = 29.208000 C 12:51:25 Temperature Measurement, Probe Number = 2, T = 28.342000 C 12:51:26 Temperature Measurement, Probe Number = 3, T = 28.760000 C 12:51:26 Temperature Measurement, Probe Number = 4, T = 28.886000 C 12:51:27 Temperature Measurement, Probe Number = 5, T = 25.561000 C 12:51:27 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.44 P3 = RTD Outside, Value = 29.21 P4 = Supply Water, Value = 28.34 P5 = Return Water, Value = 28.76 P6 = @ RTD Outside, Value = 28.89 P7 = Ambient, Value = 25.56 12:51:27 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000030, Bz = -0.000070 T 12:51:35 Move Stage, position = 0.090000 12:51:36 Stage Position Measurement, pos = 0.0900800 m 12:51:37 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:51:38 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:51:38 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:51:39 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:51:39 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:51:40 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:51:40 Main Volt Voltage Measurement, V = 0.000002 V 12:51:41 Temperature Measurement, Probe Number = 0, T = 29.445000 C 12:51:45 Temperature Measurement, Probe Number = 1, T = 29.208000 C 12:51:45 Temperature Measurement, Probe Number = 2, T = 28.345000 C 12:51:46 Temperature Measurement, Probe Number = 3, T = 28.760000 C 12:51:46 Temperature Measurement, Probe Number = 4, T = 28.892000 C 12:51:47 Temperature Measurement, Probe Number = 5, T = 25.598000 C 12:51:47 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.45 P3 = RTD Outside, Value = 29.21 P4 = Supply Water, Value = 28.34 P5 = Return Water, Value = 28.76 P6 = @ RTD Outside, Value = 28.89 P7 = Ambient, Value = 25.60 12:51:47 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000030, Bz = -0.000080 T 12:51:55 Move Stage, position = 0.092000 12:51:56 Stage Position Measurement, pos = 0.0920750 m 12:51:57 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:51:58 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:51:58 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:51:59 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:51:59 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:52:00 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:52:00 Main Volt Voltage Measurement, V = 0.000001 V 12:52:01 Temperature Measurement, Probe Number = 0, T = 29.445000 C 12:52:05 Temperature Measurement, Probe Number = 1, T = 29.212000 C 12:52:05 Temperature Measurement, Probe Number = 2, T = 28.380000 C 12:52:06 Temperature Measurement, Probe Number = 3, T = 28.767000 C 12:52:06 Temperature Measurement, Probe Number = 4, T = 28.896000 C 12:52:07 Temperature Measurement, Probe Number = 5, T = 25.719000 C 12:52:07 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.45 P3 = RTD Outside, Value = 29.21 P4 = Supply Water, Value = 28.38 P5 = Return Water, Value = 28.77 P6 = @ RTD Outside, Value = 28.90 P7 = Ambient, Value = 25.72 12:52:07 3 Axis Hall Probe Field Measurement, Bx = -0.000010, By = 0.000020, Bz = -0.000080 T 12:52:15 Move Stage, position = 0.094000 12:52:16 Stage Position Measurement, pos = 0.0940900 m 12:52:17 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:52:18 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:52:19 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:52:19 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:52:19 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:52:20 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:52:21 Main Volt Voltage Measurement, V = 0.000002 V 12:52:21 Temperature Measurement, Probe Number = 0, T = 29.445000 C 12:52:25 Temperature Measurement, Probe Number = 1, T = 29.210000 C 12:52:25 Temperature Measurement, Probe Number = 2, T = 28.436000 C 12:52:26 Temperature Measurement, Probe Number = 3, T = 28.777000 C 12:52:27 Temperature Measurement, Probe Number = 4, T = 28.900000 C 12:52:27 Temperature Measurement, Probe Number = 5, T = 25.741000 C 12:52:27 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.45 P3 = RTD Outside, Value = 29.21 P4 = Supply Water, Value = 28.44 P5 = Return Water, Value = 28.78 P6 = @ RTD Outside, Value = 28.90 P7 = Ambient, Value = 25.74 12:52:27 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000020, Bz = -0.000060 T 12:52:35 Move Stage, position = 0.096000 12:52:36 Stage Position Measurement, pos = 0.0961000 m 12:52:37 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:52:38 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:52:39 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:52:39 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:52:39 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:52:40 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:52:41 Main Volt Voltage Measurement, V = 0.000001 V 12:52:41 Temperature Measurement, Probe Number = 0, T = 29.446000 C 12:52:45 Temperature Measurement, Probe Number = 1, T = 29.212000 C 12:52:46 Temperature Measurement, Probe Number = 2, T = 28.482000 C 12:52:46 Temperature Measurement, Probe Number = 3, T = 28.780000 C 12:52:47 Temperature Measurement, Probe Number = 4, T = 28.902000 C 12:52:47 Temperature Measurement, Probe Number = 5, T = 25.756000 C 12:52:47 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.45 P3 = RTD Outside, Value = 29.21 P4 = Supply Water, Value = 28.48 P5 = Return Water, Value = 28.78 P6 = @ RTD Outside, Value = 28.90 P7 = Ambient, Value = 25.76 12:52:47 3 Axis Hall Probe Field Measurement, Bx = -0.000010, By = 0.000030, Bz = -0.000070 T 12:52:55 Move Stage, position = 0.098000 12:52:57 Stage Position Measurement, pos = 0.0981300 m 12:52:57 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:52:58 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:52:59 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:53:00 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:53:00 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:53:00 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:53:01 Main Volt Voltage Measurement, V = 0.000001 V 12:53:01 Temperature Measurement, Probe Number = 0, T = 29.452000 C 12:53:05 Temperature Measurement, Probe Number = 1, T = 29.216000 C 12:53:06 Temperature Measurement, Probe Number = 2, T = 28.513000 C 12:53:06 Temperature Measurement, Probe Number = 3, T = 28.786000 C 12:53:07 Temperature Measurement, Probe Number = 4, T = 28.906000 C 12:53:07 Temperature Measurement, Probe Number = 5, T = 25.716000 C 12:53:07 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.45 P3 = RTD Outside, Value = 29.22 P4 = Supply Water, Value = 28.51 P5 = Return Water, Value = 28.79 P6 = @ RTD Outside, Value = 28.91 P7 = Ambient, Value = 25.72 12:53:07 3 Axis Hall Probe Field Measurement, Bx = -0.000010, By = 0.000020, Bz = -0.000060 T 12:53:15 Move Stage, position = 0.100000 12:53:17 Stage Position Measurement, pos = 0.1001150 m 12:53:17 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:53:18 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:53:19 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:53:20 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:53:20 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:53:20 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:53:21 Main Volt Voltage Measurement, V = 0.000002 V 12:53:22 Temperature Measurement, Probe Number = 0, T = 29.455000 C 12:53:25 Temperature Measurement, Probe Number = 1, T = 29.222000 C 12:53:26 Temperature Measurement, Probe Number = 2, T = 28.503000 C 12:53:26 Temperature Measurement, Probe Number = 3, T = 28.784000 C 12:53:27 Temperature Measurement, Probe Number = 4, T = 28.912000 C 12:53:27 Temperature Measurement, Probe Number = 5, T = 25.705000 C 12:53:27 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.45 P3 = RTD Outside, Value = 29.22 P4 = Supply Water, Value = 28.50 P5 = Return Water, Value = 28.78 P6 = @ RTD Outside, Value = 28.91 P7 = Ambient, Value = 25.70 12:53:28 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000020, Bz = -0.000060 T 12:53:35 Move Stage, position = 0.102000 12:53:37 Stage Position Measurement, pos = 0.1021000 m 12:53:38 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:53:38 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:53:39 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:53:40 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:53:40 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:53:40 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:53:41 Main Volt Voltage Measurement, V = 0.000002 V 12:53:42 Temperature Measurement, Probe Number = 0, T = 29.453000 C 12:53:45 Temperature Measurement, Probe Number = 1, T = 29.219000 C 12:53:46 Temperature Measurement, Probe Number = 2, T = 28.456000 C 12:53:46 Temperature Measurement, Probe Number = 3, T = 28.774000 C 12:53:47 Temperature Measurement, Probe Number = 4, T = 28.918000 C 12:53:48 Temperature Measurement, Probe Number = 5, T = 25.681000 C 12:53:48 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.45 P3 = RTD Outside, Value = 29.22 P4 = Supply Water, Value = 28.46 P5 = Return Water, Value = 28.77 P6 = @ RTD Outside, Value = 28.92 P7 = Ambient, Value = 25.68 12:53:48 3 Axis Hall Probe Field Measurement, Bx = -0.000010, By = 0.000020, Bz = -0.000060 T 12:53:55 Move Stage, position = 0.104000 12:53:57 Stage Position Measurement, pos = 0.1040800 m 12:53:58 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:53:58 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:53:59 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:54:00 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:54:00 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:54:01 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:54:01 Main Volt Voltage Measurement, V = 0.000001 V 12:54:02 Temperature Measurement, Probe Number = 0, T = 29.446000 C 12:54:05 Temperature Measurement, Probe Number = 1, T = 29.212000 C 12:54:06 Temperature Measurement, Probe Number = 2, T = 28.456000 C 12:54:07 Temperature Measurement, Probe Number = 3, T = 28.758000 C 12:54:07 Temperature Measurement, Probe Number = 4, T = 28.914000 C 12:54:08 Temperature Measurement, Probe Number = 5, T = 25.792000 C 12:54:08 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.45 P3 = RTD Outside, Value = 29.21 P4 = Supply Water, Value = 28.46 P5 = Return Water, Value = 28.76 P6 = @ RTD Outside, Value = 28.91 P7 = Ambient, Value = 25.79 12:54:08 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000030, Bz = -0.000050 T 12:54:15 Move Stage, position = 0.106000 12:54:17 Stage Position Measurement, pos = 0.1060850 m 12:54:18 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:54:19 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:54:19 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:54:20 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:54:20 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:54:21 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:54:21 Main Volt Voltage Measurement, V = 0.000001 V 12:54:22 Temperature Measurement, Probe Number = 0, T = 29.445000 C 12:54:25 Temperature Measurement, Probe Number = 1, T = 29.208000 C 12:54:26 Temperature Measurement, Probe Number = 2, T = 28.478000 C 12:54:27 Temperature Measurement, Probe Number = 3, T = 28.756000 C 12:54:27 Temperature Measurement, Probe Number = 4, T = 28.902000 C 12:54:28 Temperature Measurement, Probe Number = 5, T = 25.969000 C 12:54:28 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.45 P3 = RTD Outside, Value = 29.21 P4 = Supply Water, Value = 28.48 P5 = Return Water, Value = 28.76 P6 = @ RTD Outside, Value = 28.90 P7 = Ambient, Value = 25.97 12:54:28 3 Axis Hall Probe Field Measurement, Bx = -0.000020, By = 0.000030, Bz = -0.000060 T 12:54:36 Move Stage, position = 0.108000 12:54:37 Stage Position Measurement, pos = 0.1080950 m 12:54:38 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:54:39 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:54:39 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:54:40 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:54:40 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:54:41 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:54:41 Main Volt Voltage Measurement, V = 0.000001 V 12:54:42 Temperature Measurement, Probe Number = 0, T = 29.448000 C 12:54:46 Temperature Measurement, Probe Number = 1, T = 29.205000 C 12:54:46 Temperature Measurement, Probe Number = 2, T = 28.482000 C 12:54:47 Temperature Measurement, Probe Number = 3, T = 28.773000 C 12:54:47 Temperature Measurement, Probe Number = 4, T = 28.885000 C 12:54:48 Temperature Measurement, Probe Number = 5, T = 26.134000 C 12:54:48 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.45 P3 = RTD Outside, Value = 29.20 P4 = Supply Water, Value = 28.48 P5 = Return Water, Value = 28.77 P6 = @ RTD Outside, Value = 28.89 P7 = Ambient, Value = 26.13 12:54:48 3 Axis Hall Probe Field Measurement, Bx = -0.000010, By = 0.000020, Bz = -0.000040 T 12:54:56 Move Stage, position = 0.110000 12:54:57 Stage Position Measurement, pos = 0.1101050 m 12:54:58 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:54:59 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:54:59 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:55:00 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:55:00 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:55:01 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:55:01 Main Volt Voltage Measurement, V = 0.000001 V 12:55:02 Temperature Measurement, Probe Number = 0, T = 29.450000 C 12:55:06 Temperature Measurement, Probe Number = 1, T = 29.192000 C 12:55:06 Temperature Measurement, Probe Number = 2, T = 28.499000 C 12:55:07 Temperature Measurement, Probe Number = 3, T = 28.786000 C 12:55:07 Temperature Measurement, Probe Number = 4, T = 28.867000 C 12:55:08 Temperature Measurement, Probe Number = 5, T = 26.081000 C 12:55:08 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.45 P3 = RTD Outside, Value = 29.19 P4 = Supply Water, Value = 28.50 P5 = Return Water, Value = 28.79 P6 = @ RTD Outside, Value = 28.87 P7 = Ambient, Value = 26.08 12:55:08 3 Axis Hall Probe Field Measurement, Bx = -0.000010, By = 0.000020, Bz = -0.000040 T 12:55:16 Move Stage, position = 0.112000 12:55:17 Stage Position Measurement, pos = 0.1121100 m 12:55:18 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:55:19 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:55:20 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:55:20 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:55:20 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:55:21 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:55:22 Main Volt Voltage Measurement, V = 0.000001 V 12:55:22 Temperature Measurement, Probe Number = 0, T = 29.446000 C 12:55:26 Temperature Measurement, Probe Number = 1, T = 29.179000 C 12:55:26 Temperature Measurement, Probe Number = 2, T = 28.488000 C 12:55:27 Temperature Measurement, Probe Number = 3, T = 28.789000 C 12:55:28 Temperature Measurement, Probe Number = 4, T = 28.844000 C 12:55:28 Temperature Measurement, Probe Number = 5, T = 25.956000 C 12:55:28 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.45 P3 = RTD Outside, Value = 29.18 P4 = Supply Water, Value = 28.49 P5 = Return Water, Value = 28.79 P6 = @ RTD Outside, Value = 28.84 P7 = Ambient, Value = 25.96 12:55:28 3 Axis Hall Probe Field Measurement, Bx = -0.000010, By = 0.000020, Bz = -0.000040 T 12:55:36 Move Stage, position = 0.114000 12:55:38 Stage Position Measurement, pos = 0.1141100 m 12:55:38 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:55:39 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:55:40 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:55:40 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:55:40 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:55:41 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:55:42 Main Volt Voltage Measurement, V = 0.000001 V 12:55:42 Temperature Measurement, Probe Number = 0, T = 29.443000 C 12:55:46 Temperature Measurement, Probe Number = 1, T = 29.187000 C 12:55:47 Temperature Measurement, Probe Number = 2, T = 28.485000 C 12:55:47 Temperature Measurement, Probe Number = 3, T = 28.780000 C 12:55:48 Temperature Measurement, Probe Number = 4, T = 28.847000 C 12:55:48 Temperature Measurement, Probe Number = 5, T = 25.913000 C 12:55:48 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.44 P3 = RTD Outside, Value = 29.19 P4 = Supply Water, Value = 28.48 P5 = Return Water, Value = 28.78 P6 = @ RTD Outside, Value = 28.85 P7 = Ambient, Value = 25.91 12:55:48 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000030, Bz = -0.000030 T 12:55:56 Move Stage, position = 0.116000 12:55:58 Stage Position Measurement, pos = 0.1161550 m 12:55:58 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:55:59 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:56:00 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:56:01 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:56:01 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:56:01 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:56:02 Main Volt Voltage Measurement, V = 0.000001 V 12:56:02 Temperature Measurement, Probe Number = 0, T = 29.443000 C 12:56:06 Temperature Measurement, Probe Number = 1, T = 29.185000 C 12:56:07 Temperature Measurement, Probe Number = 2, T = 28.478000 C 12:56:07 Temperature Measurement, Probe Number = 3, T = 28.777000 C 12:56:08 Temperature Measurement, Probe Number = 4, T = 28.849000 C 12:56:08 Temperature Measurement, Probe Number = 5, T = 25.888000 C 12:56:08 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.44 P3 = RTD Outside, Value = 29.18 P4 = Supply Water, Value = 28.48 P5 = Return Water, Value = 28.78 P6 = @ RTD Outside, Value = 28.85 P7 = Ambient, Value = 25.89 12:56:08 3 Axis Hall Probe Field Measurement, Bx = -0.000010, By = 0.000020, Bz = -0.000030 T 12:56:16 Move Stage, position = 0.118000 12:56:18 Stage Position Measurement, pos = 0.1181350 m 12:56:18 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:56:19 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:56:20 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:56:21 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:56:21 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:56:21 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:56:22 Main Volt Voltage Measurement, V = 0.000001 V 12:56:22 Temperature Measurement, Probe Number = 0, T = 29.448000 C 12:56:26 Temperature Measurement, Probe Number = 1, T = 29.200000 C 12:56:27 Temperature Measurement, Probe Number = 2, T = 28.477000 C 12:56:27 Temperature Measurement, Probe Number = 3, T = 28.778000 C 12:56:28 Temperature Measurement, Probe Number = 4, T = 28.851000 C 12:56:28 Temperature Measurement, Probe Number = 5, T = 25.904000 C 12:56:28 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.45 P3 = RTD Outside, Value = 29.20 P4 = Supply Water, Value = 28.48 P5 = Return Water, Value = 28.78 P6 = @ RTD Outside, Value = 28.85 P7 = Ambient, Value = 25.90 12:56:29 3 Axis Hall Probe Field Measurement, Bx = -0.000010, By = 0.000030, Bz = -0.000030 T 12:56:36 Move Stage, position = 0.120000 12:56:38 Stage Position Measurement, pos = 0.1201150 m 12:56:39 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:56:39 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:56:40 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:56:41 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:56:41 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:56:41 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:56:42 Main Volt Voltage Measurement, V = 0.000001 V 12:56:43 Temperature Measurement, Probe Number = 0, T = 29.452000 C 12:56:46 Temperature Measurement, Probe Number = 1, T = 29.212000 C 12:56:47 Temperature Measurement, Probe Number = 2, T = 28.506000 C 12:56:47 Temperature Measurement, Probe Number = 3, T = 28.778000 C 12:56:48 Temperature Measurement, Probe Number = 4, T = 28.865000 C 12:56:49 Temperature Measurement, Probe Number = 5, T = 25.903000 C 12:56:49 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.45 P3 = RTD Outside, Value = 29.21 P4 = Supply Water, Value = 28.51 P5 = Return Water, Value = 28.78 P6 = @ RTD Outside, Value = 28.86 P7 = Ambient, Value = 25.90 12:56:49 3 Axis Hall Probe Field Measurement, Bx = -0.000010, By = 0.000030, Bz = -0.000020 T 12:56:56 Move Stage, position = 0.122000 12:56:58 Stage Position Measurement, pos = 0.1221050 m 12:56:59 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:56:59 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:57:00 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:57:01 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:57:01 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:57:02 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:57:02 Main Volt Voltage Measurement, V = 0.000002 V 12:57:03 Temperature Measurement, Probe Number = 0, T = 29.454000 C 12:57:06 Temperature Measurement, Probe Number = 1, T = 29.221000 C 12:57:07 Temperature Measurement, Probe Number = 2, T = 28.526000 C 12:57:08 Temperature Measurement, Probe Number = 3, T = 28.784000 C 12:57:08 Temperature Measurement, Probe Number = 4, T = 28.883000 C 12:57:09 Temperature Measurement, Probe Number = 5, T = 25.881000 C 12:57:09 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.45 P3 = RTD Outside, Value = 29.22 P4 = Supply Water, Value = 28.53 P5 = Return Water, Value = 28.78 P6 = @ RTD Outside, Value = 28.88 P7 = Ambient, Value = 25.88 12:57:09 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000030, Bz = -0.000020 T 12:57:16 Move Stage, position = 0.124000 12:57:18 Stage Position Measurement, pos = 0.1240900 m 12:57:19 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:57:20 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:57:20 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:57:21 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:57:21 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:57:22 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:57:22 Main Volt Voltage Measurement, V = 0.000002 V 12:57:23 Temperature Measurement, Probe Number = 0, T = 29.456000 C 12:57:26 Temperature Measurement, Probe Number = 1, T = 29.224000 C 12:57:27 Temperature Measurement, Probe Number = 2, T = 28.540000 C 12:57:28 Temperature Measurement, Probe Number = 3, T = 28.789000 C 12:57:28 Temperature Measurement, Probe Number = 4, T = 28.894000 C 12:57:29 Temperature Measurement, Probe Number = 5, T = 25.896000 C 12:57:29 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.46 P3 = RTD Outside, Value = 29.22 P4 = Supply Water, Value = 28.54 P5 = Return Water, Value = 28.79 P6 = @ RTD Outside, Value = 28.89 P7 = Ambient, Value = 25.90 12:57:29 3 Axis Hall Probe Field Measurement, Bx = -0.000010, By = 0.000030, Bz = -0.000020 T 12:57:37 Move Stage, position = 0.126000 12:57:38 Stage Position Measurement, pos = 0.1261050 m 12:57:39 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:57:40 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:57:40 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:57:41 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:57:41 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:57:42 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:57:42 Main Volt Voltage Measurement, V = 0.000002 V 12:57:43 Temperature Measurement, Probe Number = 0, T = 29.457000 C 12:57:47 Temperature Measurement, Probe Number = 1, T = 29.223000 C 12:57:47 Temperature Measurement, Probe Number = 2, T = 28.549000 C 12:57:48 Temperature Measurement, Probe Number = 3, T = 28.792000 C 12:57:48 Temperature Measurement, Probe Number = 4, T = 28.905000 C 12:57:49 Temperature Measurement, Probe Number = 5, T = 25.873000 C 12:57:49 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.46 P3 = RTD Outside, Value = 29.22 P4 = Supply Water, Value = 28.55 P5 = Return Water, Value = 28.79 P6 = @ RTD Outside, Value = 28.91 P7 = Ambient, Value = 25.87 12:57:49 3 Axis Hall Probe Field Measurement, Bx = -0.000020, By = 0.000020, Bz = -0.000030 T 12:57:57 Move Stage, position = 0.128000 12:57:58 Stage Position Measurement, pos = 0.1281150 m 12:57:59 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:58:00 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:58:00 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:58:01 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:58:01 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:58:02 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:58:02 Main Volt Voltage Measurement, V = 0.000002 V 12:58:03 Temperature Measurement, Probe Number = 0, T = 29.459000 C 12:58:07 Temperature Measurement, Probe Number = 1, T = 29.225000 C 12:58:07 Temperature Measurement, Probe Number = 2, T = 28.546000 C 12:58:08 Temperature Measurement, Probe Number = 3, T = 28.784000 C 12:58:09 Temperature Measurement, Probe Number = 4, T = 28.915000 C 12:58:09 Temperature Measurement, Probe Number = 5, T = 25.800000 C 12:58:09 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.46 P3 = RTD Outside, Value = 29.23 P4 = Supply Water, Value = 28.55 P5 = Return Water, Value = 28.78 P6 = @ RTD Outside, Value = 28.91 P7 = Ambient, Value = 25.80 12:58:09 3 Axis Hall Probe Field Measurement, Bx = -0.000010, By = 0.000030, Bz = -0.000020 T 12:58:17 Move Stage, position = 0.130000 12:58:19 Stage Position Measurement, pos = 0.1301300 m 12:58:19 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:58:20 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:58:21 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:58:21 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:58:21 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:58:22 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:58:23 Main Volt Voltage Measurement, V = 0.000002 V 12:58:23 Temperature Measurement, Probe Number = 0, T = 29.458000 C 12:58:27 Temperature Measurement, Probe Number = 1, T = 29.227000 C 12:58:28 Temperature Measurement, Probe Number = 2, T = 28.533000 C 12:58:28 Temperature Measurement, Probe Number = 3, T = 28.782000 C 12:58:29 Temperature Measurement, Probe Number = 4, T = 28.922000 C 12:58:29 Temperature Measurement, Probe Number = 5, T = 25.796000 C 12:58:29 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.46 P3 = RTD Outside, Value = 29.23 P4 = Supply Water, Value = 28.53 P5 = Return Water, Value = 28.78 P6 = @ RTD Outside, Value = 28.92 P7 = Ambient, Value = 25.80 12:58:29 3 Axis Hall Probe Field Measurement, Bx = -0.000010, By = 0.000030, Bz = -0.000020 T 12:58:37 Move Stage, position = 0.132000 12:58:39 Stage Position Measurement, pos = 0.1321200 m 12:58:39 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:58:40 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:58:41 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:58:42 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:58:42 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:58:42 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:58:43 Main Volt Voltage Measurement, V = 0.000002 V 12:58:43 Temperature Measurement, Probe Number = 0, T = 29.456000 C 12:58:47 Temperature Measurement, Probe Number = 1, T = 29.229000 C 12:58:48 Temperature Measurement, Probe Number = 2, T = 28.542000 C 12:58:48 Temperature Measurement, Probe Number = 3, T = 28.785000 C 12:58:49 Temperature Measurement, Probe Number = 4, T = 28.927000 C 12:58:49 Temperature Measurement, Probe Number = 5, T = 25.793000 C 12:58:49 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.46 P3 = RTD Outside, Value = 29.23 P4 = Supply Water, Value = 28.54 P5 = Return Water, Value = 28.79 P6 = @ RTD Outside, Value = 28.93 P7 = Ambient, Value = 25.79 12:58:49 3 Axis Hall Probe Field Measurement, Bx = -0.000010, By = 0.000040, Bz = -0.000020 T 12:58:57 Move Stage, position = 0.134000 12:58:59 Stage Position Measurement, pos = 0.1341550 m 12:58:59 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:59:00 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:59:01 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:59:02 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:59:02 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:59:02 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:59:03 Main Volt Voltage Measurement, V = 0.000002 V 12:59:04 Temperature Measurement, Probe Number = 0, T = 29.455000 C 12:59:07 Temperature Measurement, Probe Number = 1, T = 29.227000 C 12:59:08 Temperature Measurement, Probe Number = 2, T = 28.544000 C 12:59:08 Temperature Measurement, Probe Number = 3, T = 28.772000 C 12:59:09 Temperature Measurement, Probe Number = 4, T = 28.930000 C 12:59:09 Temperature Measurement, Probe Number = 5, T = 25.708000 C 12:59:09 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.45 P3 = RTD Outside, Value = 29.23 P4 = Supply Water, Value = 28.54 P5 = Return Water, Value = 28.77 P6 = @ RTD Outside, Value = 28.93 P7 = Ambient, Value = 25.71 12:59:10 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000040, Bz = -0.000020 T 12:59:17 Move Stage, position = 0.136000 12:59:19 Stage Position Measurement, pos = 0.1361600 m 12:59:20 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:59:20 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:59:21 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:59:22 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:59:22 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:59:22 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:59:23 Main Volt Voltage Measurement, V = 0.000002 V 12:59:24 Temperature Measurement, Probe Number = 0, T = 29.452000 C 12:59:27 Temperature Measurement, Probe Number = 1, T = 29.223000 C 12:59:28 Temperature Measurement, Probe Number = 2, T = 28.525000 C 12:59:28 Temperature Measurement, Probe Number = 3, T = 28.742000 C 12:59:29 Temperature Measurement, Probe Number = 4, T = 28.931000 C 12:59:30 Temperature Measurement, Probe Number = 5, T = 25.727000 C 12:59:30 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.45 P3 = RTD Outside, Value = 29.22 P4 = Supply Water, Value = 28.52 P5 = Return Water, Value = 28.74 P6 = @ RTD Outside, Value = 28.93 P7 = Ambient, Value = 25.73 12:59:30 3 Axis Hall Probe Field Measurement, Bx = -0.000010, By = 0.000030, Bz = -0.000010 T 12:59:37 Move Stage, position = 0.138000 12:59:39 Stage Position Measurement, pos = 0.1381350 m 12:59:40 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:59:40 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:59:41 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:59:42 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:59:42 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 12:59:43 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 12:59:43 Main Volt Voltage Measurement, V = 0.000002 V 12:59:44 Temperature Measurement, Probe Number = 0, T = 29.450000 C 12:59:47 Temperature Measurement, Probe Number = 1, T = 29.222000 C 12:59:48 Temperature Measurement, Probe Number = 2, T = 28.515000 C 12:59:49 Temperature Measurement, Probe Number = 3, T = 28.707000 C 12:59:49 Temperature Measurement, Probe Number = 4, T = 28.932000 C 12:59:50 Temperature Measurement, Probe Number = 5, T = 25.625000 C 12:59:50 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.45 P3 = RTD Outside, Value = 29.22 P4 = Supply Water, Value = 28.52 P5 = Return Water, Value = 28.71 P6 = @ RTD Outside, Value = 28.93 P7 = Ambient, Value = 25.62 12:59:50 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000030, Bz = -0.000010 T 12:59:57 Move Stage, position = 0.140000 12:59:59 Stage Position Measurement, pos = 0.1401250 m 13:00:00 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:00:01 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:00:01 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:00:02 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:00:02 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 13:00:03 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:00:03 Main Volt Voltage Measurement, V = 0.000002 V 13:00:04 Temperature Measurement, Probe Number = 0, T = 29.447000 C 13:00:07 Temperature Measurement, Probe Number = 1, T = 29.221000 C 13:00:08 Temperature Measurement, Probe Number = 2, T = 28.501000 C 13:00:09 Temperature Measurement, Probe Number = 3, T = 28.684000 C 13:00:09 Temperature Measurement, Probe Number = 4, T = 28.932000 C 13:00:10 Temperature Measurement, Probe Number = 5, T = 25.610000 C 13:00:10 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.45 P3 = RTD Outside, Value = 29.22 P4 = Supply Water, Value = 28.50 P5 = Return Water, Value = 28.68 P6 = @ RTD Outside, Value = 28.93 P7 = Ambient, Value = 25.61 13:00:10 3 Axis Hall Probe Field Measurement, Bx = -0.000010, By = 0.000030, Bz = -0.000010 T 13:00:18 Move Stage, position = 0.142000 13:00:19 Stage Position Measurement, pos = 0.1421100 m 13:00:20 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:00:21 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:00:21 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:00:22 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:00:22 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 13:00:23 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:00:23 Main Volt Voltage Measurement, V = 0.000002 V 13:00:24 Temperature Measurement, Probe Number = 0, T = 29.445000 C 13:00:28 Temperature Measurement, Probe Number = 1, T = 29.220000 C 13:00:28 Temperature Measurement, Probe Number = 2, T = 28.488000 C 13:00:29 Temperature Measurement, Probe Number = 3, T = 28.689000 C 13:00:29 Temperature Measurement, Probe Number = 4, T = 28.933000 C 13:00:30 Temperature Measurement, Probe Number = 5, T = 25.608000 C 13:00:30 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.45 P3 = RTD Outside, Value = 29.22 P4 = Supply Water, Value = 28.49 P5 = Return Water, Value = 28.69 P6 = @ RTD Outside, Value = 28.93 P7 = Ambient, Value = 25.61 13:00:30 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000020, Bz = -0.000020 T 13:00:38 Move Stage, position = 0.144000 13:00:39 Stage Position Measurement, pos = 0.1441100 m 13:00:40 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:00:41 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:00:41 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:00:42 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:00:42 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 13:00:43 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:00:43 Main Volt Voltage Measurement, V = 0.000002 V 13:00:44 Temperature Measurement, Probe Number = 0, T = 29.444000 C 13:00:48 Temperature Measurement, Probe Number = 1, T = 29.222000 C 13:00:48 Temperature Measurement, Probe Number = 2, T = 28.466000 C 13:00:49 Temperature Measurement, Probe Number = 3, T = 28.693000 C 13:00:49 Temperature Measurement, Probe Number = 4, T = 28.934000 C 13:00:50 Temperature Measurement, Probe Number = 5, T = 25.692000 C 13:00:50 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.44 P3 = RTD Outside, Value = 29.22 P4 = Supply Water, Value = 28.47 P5 = Return Water, Value = 28.69 P6 = @ RTD Outside, Value = 28.93 P7 = Ambient, Value = 25.69 13:00:50 3 Axis Hall Probe Field Measurement, Bx = -0.000010, By = 0.000040, Bz = -0.000020 T 13:00:58 Move Stage, position = 0.146000 13:00:59 Stage Position Measurement, pos = 0.1461400 m 13:01:00 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:01:01 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:01:02 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:01:02 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:01:02 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 13:01:03 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:01:04 Main Volt Voltage Measurement, V = 0.000002 V 13:01:04 Temperature Measurement, Probe Number = 0, T = 29.445000 C 13:01:08 Temperature Measurement, Probe Number = 1, T = 29.224000 C 13:01:08 Temperature Measurement, Probe Number = 2, T = 28.432000 C 13:01:09 Temperature Measurement, Probe Number = 3, T = 28.704000 C 13:01:10 Temperature Measurement, Probe Number = 4, T = 28.935000 C 13:01:10 Temperature Measurement, Probe Number = 5, T = 25.701000 C 13:01:10 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.45 P3 = RTD Outside, Value = 29.22 P4 = Supply Water, Value = 28.43 P5 = Return Water, Value = 28.70 P6 = @ RTD Outside, Value = 28.93 P7 = Ambient, Value = 25.70 13:01:10 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000030, Bz = -0.000020 T 13:01:18 Move Stage, position = 0.148000 13:01:20 Stage Position Measurement, pos = 0.1481550 m 13:01:20 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:01:21 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:01:22 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:01:22 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:01:22 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 13:01:23 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:01:24 Main Volt Voltage Measurement, V = 0.000002 V 13:01:24 Temperature Measurement, Probe Number = 0, T = 29.446000 C 13:01:28 Temperature Measurement, Probe Number = 1, T = 29.211000 C 13:01:29 Temperature Measurement, Probe Number = 2, T = 28.382000 C 13:01:29 Temperature Measurement, Probe Number = 3, T = 28.711000 C 13:01:30 Temperature Measurement, Probe Number = 4, T = 28.931000 C 13:01:30 Temperature Measurement, Probe Number = 5, T = 25.705000 C 13:01:30 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.45 P3 = RTD Outside, Value = 29.21 P4 = Supply Water, Value = 28.38 P5 = Return Water, Value = 28.71 P6 = @ RTD Outside, Value = 28.93 P7 = Ambient, Value = 25.70 13:01:30 3 Axis Hall Probe Field Measurement, Bx = -0.000010, By = 0.000040, Bz = -0.000020 T 13:01:38 Move Stage, position = 0.150000 13:01:40 Stage Position Measurement, pos = 0.1501600 m 13:01:40 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:01:41 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:01:42 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:01:43 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:01:43 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 13:01:43 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:01:44 Main Volt Voltage Measurement, V = 0.000002 V 13:01:44 Temperature Measurement, Probe Number = 0, T = 29.447000 C 13:01:48 Temperature Measurement, Probe Number = 1, T = 29.212000 C 13:01:49 Temperature Measurement, Probe Number = 2, T = 28.343000 C 13:01:49 Temperature Measurement, Probe Number = 3, T = 28.705000 C 13:01:50 Temperature Measurement, Probe Number = 4, T = 28.922000 C 13:01:50 Temperature Measurement, Probe Number = 5, T = 25.722000 C 13:01:50 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.45 P3 = RTD Outside, Value = 29.21 P4 = Supply Water, Value = 28.34 P5 = Return Water, Value = 28.70 P6 = @ RTD Outside, Value = 28.92 P7 = Ambient, Value = 25.72 13:01:50 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000020, Bz = -0.000020 T 13:01:58 Move Stage, position = 0.152000 13:02:00 Stage Position Measurement, pos = 0.1521850 m 13:02:00 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:02:01 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:02:02 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:02:03 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:02:03 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 13:02:03 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:02:04 Main Volt Voltage Measurement, V = 0.000002 V 13:02:05 Temperature Measurement, Probe Number = 0, T = 29.452000 C 13:02:08 Temperature Measurement, Probe Number = 1, T = 29.220000 C 13:02:09 Temperature Measurement, Probe Number = 2, T = 28.352000 C 13:02:09 Temperature Measurement, Probe Number = 3, T = 28.716000 C 13:02:10 Temperature Measurement, Probe Number = 4, T = 28.925000 C 13:02:10 Temperature Measurement, Probe Number = 5, T = 25.787000 C 13:02:10 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.45 P3 = RTD Outside, Value = 29.22 P4 = Supply Water, Value = 28.35 P5 = Return Water, Value = 28.72 P6 = @ RTD Outside, Value = 28.93 P7 = Ambient, Value = 25.79 13:02:11 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000030, Bz = -0.000010 T 13:02:18 Move Stage, position = 0.154000 13:02:20 Stage Position Measurement, pos = 0.1541550 m 13:02:21 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:02:21 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:02:22 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:02:23 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:02:23 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 13:02:23 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:02:24 Main Volt Voltage Measurement, V = 0.000002 V 13:02:25 Temperature Measurement, Probe Number = 0, T = 29.453000 C 13:02:28 Temperature Measurement, Probe Number = 1, T = 29.223000 C 13:02:29 Temperature Measurement, Probe Number = 2, T = 28.370000 C 13:02:29 Temperature Measurement, Probe Number = 3, T = 28.711000 C 13:02:30 Temperature Measurement, Probe Number = 4, T = 28.928000 C 13:02:31 Temperature Measurement, Probe Number = 5, T = 25.688000 C 13:02:31 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.45 P3 = RTD Outside, Value = 29.22 P4 = Supply Water, Value = 28.37 P5 = Return Water, Value = 28.71 P6 = @ RTD Outside, Value = 28.93 P7 = Ambient, Value = 25.69 13:02:31 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000030, Bz = -0.000020 T 13:02:38 Move Stage, position = 0.156000 13:02:40 Stage Position Measurement, pos = 0.1561650 m 13:02:41 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:02:41 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:02:42 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:02:43 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:02:43 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 13:02:44 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:02:44 Main Volt Voltage Measurement, V = 0.000002 V 13:02:45 Temperature Measurement, Probe Number = 0, T = 29.452000 C 13:02:48 Temperature Measurement, Probe Number = 1, T = 29.223000 C 13:02:49 Temperature Measurement, Probe Number = 2, T = 28.394000 C 13:02:50 Temperature Measurement, Probe Number = 3, T = 28.677000 C 13:02:50 Temperature Measurement, Probe Number = 4, T = 28.930000 C 13:02:51 Temperature Measurement, Probe Number = 5, T = 25.683000 C 13:02:51 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.45 P3 = RTD Outside, Value = 29.22 P4 = Supply Water, Value = 28.39 P5 = Return Water, Value = 28.68 P6 = @ RTD Outside, Value = 28.93 P7 = Ambient, Value = 25.68 13:02:51 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000030, Bz = -0.000020 T 13:02:58 Move Stage, position = 0.158000 13:03:00 Stage Position Measurement, pos = 0.1581600 m 13:03:01 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:03:02 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:03:02 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:03:03 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:03:03 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 13:03:04 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:03:04 Main Volt Voltage Measurement, V = 0.000002 V 13:03:05 Temperature Measurement, Probe Number = 0, T = 29.451000 C 13:03:08 Temperature Measurement, Probe Number = 1, T = 29.218000 C 13:03:09 Temperature Measurement, Probe Number = 2, T = 28.420000 C 13:03:10 Temperature Measurement, Probe Number = 3, T = 28.638000 C 13:03:10 Temperature Measurement, Probe Number = 4, T = 28.929000 C 13:03:11 Temperature Measurement, Probe Number = 5, T = 25.809000 C 13:03:11 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.45 P3 = RTD Outside, Value = 29.22 P4 = Supply Water, Value = 28.42 P5 = Return Water, Value = 28.64 P6 = @ RTD Outside, Value = 28.93 P7 = Ambient, Value = 25.81 13:03:11 3 Axis Hall Probe Field Measurement, Bx = -0.000010, By = 0.000030, Bz = -0.000020 T 13:03:19 Move Stage, position = 0.160000 13:03:20 Stage Position Measurement, pos = 0.1601500 m 13:03:21 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:03:22 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:03:22 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:03:23 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:03:23 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 13:03:24 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:03:24 Main Volt Voltage Measurement, V = 0.000002 V 13:03:25 Temperature Measurement, Probe Number = 0, T = 29.451000 C 13:03:29 Temperature Measurement, Probe Number = 1, T = 29.221000 C 13:03:29 Temperature Measurement, Probe Number = 2, T = 28.397000 C 13:03:30 Temperature Measurement, Probe Number = 3, T = 28.645000 C 13:03:30 Temperature Measurement, Probe Number = 4, T = 28.930000 C 13:03:31 Temperature Measurement, Probe Number = 5, T = 25.832000 C 13:03:31 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.45 P3 = RTD Outside, Value = 29.22 P4 = Supply Water, Value = 28.40 P5 = Return Water, Value = 28.64 P6 = @ RTD Outside, Value = 28.93 P7 = Ambient, Value = 25.83 13:03:31 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000040, Bz = -0.000020 T 13:03:39 Move Stage, position = 0.162000 13:03:40 Stage Position Measurement, pos = 0.1621450 m 13:03:41 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:03:42 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:03:42 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:03:43 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:03:43 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 13:03:44 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:03:44 Main Volt Voltage Measurement, V = 0.000002 V 13:03:45 Temperature Measurement, Probe Number = 0, T = 29.452000 C 13:03:49 Temperature Measurement, Probe Number = 1, T = 29.220000 C 13:03:49 Temperature Measurement, Probe Number = 2, T = 28.397000 C 13:03:50 Temperature Measurement, Probe Number = 3, T = 28.675000 C 13:03:50 Temperature Measurement, Probe Number = 4, T = 28.930000 C 13:03:51 Temperature Measurement, Probe Number = 5, T = 25.816000 C 13:03:51 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.45 P3 = RTD Outside, Value = 29.22 P4 = Supply Water, Value = 28.40 P5 = Return Water, Value = 28.68 P6 = @ RTD Outside, Value = 28.93 P7 = Ambient, Value = 25.82 13:03:51 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000030, Bz = -0.000020 T 13:03:59 Move Stage, position = 0.164000 13:04:00 Stage Position Measurement, pos = 0.1641550 m 13:04:01 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:04:02 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:04:03 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:04:03 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:04:03 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 13:04:04 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:04:05 Main Volt Voltage Measurement, V = 0.000002 V 13:04:05 Temperature Measurement, Probe Number = 0, T = 29.454000 C 13:04:09 Temperature Measurement, Probe Number = 1, T = 29.218000 C 13:04:09 Temperature Measurement, Probe Number = 2, T = 28.402000 C 13:04:10 Temperature Measurement, Probe Number = 3, T = 28.708000 C 13:04:11 Temperature Measurement, Probe Number = 4, T = 28.925000 C 13:04:11 Temperature Measurement, Probe Number = 5, T = 25.819000 C 13:04:11 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.45 P3 = RTD Outside, Value = 29.22 P4 = Supply Water, Value = 28.40 P5 = Return Water, Value = 28.71 P6 = @ RTD Outside, Value = 28.93 P7 = Ambient, Value = 25.82 13:04:11 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000040, Bz = -0.000020 T 13:04:19 Move Stage, position = 0.166000 13:04:21 Stage Position Measurement, pos = 0.1661850 m 13:04:21 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:04:22 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:04:23 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:04:23 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:04:23 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 13:04:24 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:04:25 Main Volt Voltage Measurement, V = 0.000002 V 13:04:25 Temperature Measurement, Probe Number = 0, T = 29.456000 C 13:04:29 Temperature Measurement, Probe Number = 1, T = 29.222000 C 13:04:30 Temperature Measurement, Probe Number = 2, T = 28.399000 C 13:04:30 Temperature Measurement, Probe Number = 3, T = 28.724000 C 13:04:31 Temperature Measurement, Probe Number = 4, T = 28.925000 C 13:04:31 Temperature Measurement, Probe Number = 5, T = 25.834000 C 13:04:31 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.46 P3 = RTD Outside, Value = 29.22 P4 = Supply Water, Value = 28.40 P5 = Return Water, Value = 28.72 P6 = @ RTD Outside, Value = 28.93 P7 = Ambient, Value = 25.83 13:04:31 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000030, Bz = -0.000010 T 13:04:39 Move Stage, position = 0.168000 13:04:41 Stage Position Measurement, pos = 0.1681800 m 13:04:41 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:04:42 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:04:43 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:04:44 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:04:44 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 13:04:44 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:04:45 Main Volt Voltage Measurement, V = 0.000002 V 13:04:45 Temperature Measurement, Probe Number = 0, T = 29.457000 C 13:04:49 Temperature Measurement, Probe Number = 1, T = 29.218000 C 13:04:50 Temperature Measurement, Probe Number = 2, T = 28.446000 C 13:04:50 Temperature Measurement, Probe Number = 3, T = 28.758000 C 13:04:51 Temperature Measurement, Probe Number = 4, T = 28.923000 C 13:04:51 Temperature Measurement, Probe Number = 5, T = 25.916000 C 13:04:51 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.46 P3 = RTD Outside, Value = 29.22 P4 = Supply Water, Value = 28.45 P5 = Return Water, Value = 28.76 P6 = @ RTD Outside, Value = 28.92 P7 = Ambient, Value = 25.92 13:04:51 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000030, Bz = -0.000020 T 13:04:59 Move Stage, position = 0.170000 13:05:01 Stage Position Measurement, pos = 0.1702100 m 13:05:02 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:05:02 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:05:03 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:05:04 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:05:04 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 13:05:04 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:05:05 Main Volt Voltage Measurement, V = 0.000001 V 13:05:06 Temperature Measurement, Probe Number = 0, T = 29.457000 C 13:05:09 Temperature Measurement, Probe Number = 1, T = 29.217000 C 13:05:10 Temperature Measurement, Probe Number = 2, T = 28.504000 C 13:05:10 Temperature Measurement, Probe Number = 3, T = 28.764000 C 13:05:11 Temperature Measurement, Probe Number = 4, T = 28.920000 C 13:05:12 Temperature Measurement, Probe Number = 5, T = 25.977000 C 13:05:12 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.46 P3 = RTD Outside, Value = 29.22 P4 = Supply Water, Value = 28.50 P5 = Return Water, Value = 28.76 P6 = @ RTD Outside, Value = 28.92 P7 = Ambient, Value = 25.98 13:05:12 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000030, Bz = -0.000020 T 13:05:19 Move Stage, position = 0.172000 13:05:21 Stage Position Measurement, pos = 0.1721900 m 13:05:22 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:05:22 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:05:23 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:05:24 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:05:24 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 13:05:25 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:05:25 Main Volt Voltage Measurement, V = 0.000002 V 13:05:26 Temperature Measurement, Probe Number = 0, T = 29.458000 C 13:05:29 Temperature Measurement, Probe Number = 1, T = 29.224000 C 13:05:30 Temperature Measurement, Probe Number = 2, T = 28.527000 C 13:05:30 Temperature Measurement, Probe Number = 3, T = 28.781000 C 13:05:31 Temperature Measurement, Probe Number = 4, T = 28.920000 C 13:05:32 Temperature Measurement, Probe Number = 5, T = 25.938000 C 13:05:32 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.46 P3 = RTD Outside, Value = 29.22 P4 = Supply Water, Value = 28.53 P5 = Return Water, Value = 28.78 P6 = @ RTD Outside, Value = 28.92 P7 = Ambient, Value = 25.94 13:05:32 3 Axis Hall Probe Field Measurement, Bx = -0.000010, By = 0.000040, Bz = -0.000010 T 13:05:39 Move Stage, position = 0.174000 13:05:41 Stage Position Measurement, pos = 0.1741850 m 13:05:42 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:05:42 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:05:43 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:05:44 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:05:44 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 13:05:45 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:05:45 Main Volt Voltage Measurement, V = 0.000002 V 13:05:46 Temperature Measurement, Probe Number = 0, T = 29.462000 C 13:05:49 Temperature Measurement, Probe Number = 1, T = 29.226000 C 13:05:50 Temperature Measurement, Probe Number = 2, T = 28.532000 C 13:05:51 Temperature Measurement, Probe Number = 3, T = 28.799000 C 13:05:51 Temperature Measurement, Probe Number = 4, T = 28.925000 C 13:05:52 Temperature Measurement, Probe Number = 5, T = 25.877000 C 13:05:52 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.46 P3 = RTD Outside, Value = 29.23 P4 = Supply Water, Value = 28.53 P5 = Return Water, Value = 28.80 P6 = @ RTD Outside, Value = 28.93 P7 = Ambient, Value = 25.88 13:05:52 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000030, Bz = -0.000010 T 13:06:00 Move Stage, position = 0.176000 13:06:01 Stage Position Measurement, pos = 0.1761950 m 13:06:02 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:06:03 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:06:03 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:06:04 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:06:04 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 13:06:05 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:06:05 Main Volt Voltage Measurement, V = 0.000002 V 13:06:06 Temperature Measurement, Probe Number = 0, T = 29.462000 C 13:06:10 Temperature Measurement, Probe Number = 1, T = 29.207000 C 13:06:10 Temperature Measurement, Probe Number = 2, T = 28.511000 C 13:06:11 Temperature Measurement, Probe Number = 3, T = 28.813000 C 13:06:11 Temperature Measurement, Probe Number = 4, T = 28.914000 C 13:06:12 Temperature Measurement, Probe Number = 5, T = 25.901000 C 13:06:12 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.46 P3 = RTD Outside, Value = 29.21 P4 = Supply Water, Value = 28.51 P5 = Return Water, Value = 28.81 P6 = @ RTD Outside, Value = 28.91 P7 = Ambient, Value = 25.90 13:06:12 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000020, Bz = -0.000030 T 13:06:20 Move Stage, position = 0.178000 13:06:21 Stage Position Measurement, pos = 0.1782000 m 13:06:22 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:06:23 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:06:23 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:06:24 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:06:24 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 13:06:25 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:06:25 Main Volt Voltage Measurement, V = 0.000002 V 13:06:26 Temperature Measurement, Probe Number = 0, T = 29.457000 C 13:06:30 Temperature Measurement, Probe Number = 1, T = 29.213000 C 13:06:30 Temperature Measurement, Probe Number = 2, T = 28.478000 C 13:06:31 Temperature Measurement, Probe Number = 3, T = 28.807000 C 13:06:31 Temperature Measurement, Probe Number = 4, T = 28.909000 C 13:06:32 Temperature Measurement, Probe Number = 5, T = 25.909000 C 13:06:32 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.46 P3 = RTD Outside, Value = 29.21 P4 = Supply Water, Value = 28.48 P5 = Return Water, Value = 28.81 P6 = @ RTD Outside, Value = 28.91 P7 = Ambient, Value = 25.91 13:06:32 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000040, Bz = -0.000030 T 13:06:40 Move Stage, position = 0.180000 13:06:41 Stage Position Measurement, pos = 0.1802000 m 13:06:42 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:06:43 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:06:43 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:06:44 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:06:44 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 13:06:45 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:06:46 Main Volt Voltage Measurement, V = 0.000002 V 13:06:46 Temperature Measurement, Probe Number = 0, T = 29.462000 C 13:06:50 Temperature Measurement, Probe Number = 1, T = 29.227000 C 13:06:50 Temperature Measurement, Probe Number = 2, T = 28.488000 C 13:06:51 Temperature Measurement, Probe Number = 3, T = 28.807000 C 13:06:51 Temperature Measurement, Probe Number = 4, T = 28.914000 C 13:06:52 Temperature Measurement, Probe Number = 5, T = 25.943000 C 13:06:52 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.46 P3 = RTD Outside, Value = 29.23 P4 = Supply Water, Value = 28.49 P5 = Return Water, Value = 28.81 P6 = @ RTD Outside, Value = 28.91 P7 = Ambient, Value = 25.94 13:06:52 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000040, Bz = -0.000020 T 13:07:00 Move Stage, position = 0.182000 13:07:01 Stage Position Measurement, pos = 0.1822050 m 13:07:02 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:07:03 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:07:04 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:07:04 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:07:04 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 13:07:05 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:07:06 Main Volt Voltage Measurement, V = 0.000001 V 13:07:06 Temperature Measurement, Probe Number = 0, T = 29.466000 C 13:07:10 Temperature Measurement, Probe Number = 1, T = 29.228000 C 13:07:10 Temperature Measurement, Probe Number = 2, T = 28.543000 C 13:07:11 Temperature Measurement, Probe Number = 3, T = 28.790000 C 13:07:12 Temperature Measurement, Probe Number = 4, T = 28.919000 C 13:07:12 Temperature Measurement, Probe Number = 5, T = 26.215000 C 13:07:12 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.47 P3 = RTD Outside, Value = 29.23 P4 = Supply Water, Value = 28.54 P5 = Return Water, Value = 28.79 P6 = @ RTD Outside, Value = 28.92 P7 = Ambient, Value = 26.21 13:07:12 3 Axis Hall Probe Field Measurement, Bx = 0.000010, By = 0.000040, Bz = -0.000020 T 13:07:20 Move Stage, position = 0.184000 13:07:22 Stage Position Measurement, pos = 0.1842300 m 13:07:22 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:07:23 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:07:24 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:07:24 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:07:24 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 13:07:25 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:07:26 Main Volt Voltage Measurement, V = 0.000001 V 13:07:26 Temperature Measurement, Probe Number = 0, T = 29.471000 C 13:07:30 Temperature Measurement, Probe Number = 1, T = 29.226000 C 13:07:31 Temperature Measurement, Probe Number = 2, T = 28.593000 C 13:07:31 Temperature Measurement, Probe Number = 3, T = 28.806000 C 13:07:32 Temperature Measurement, Probe Number = 4, T = 28.917000 C 13:07:32 Temperature Measurement, Probe Number = 5, T = 26.131000 C 13:07:32 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.47 P3 = RTD Outside, Value = 29.23 P4 = Supply Water, Value = 28.59 P5 = Return Water, Value = 28.81 P6 = @ RTD Outside, Value = 28.92 P7 = Ambient, Value = 26.13 13:07:32 3 Axis Hall Probe Field Measurement, Bx = -0.000010, By = 0.000030, Bz = -0.000020 T 13:07:40 Move Stage, position = 0.186000 13:07:42 Stage Position Measurement, pos = 0.1862550 m 13:07:42 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:07:43 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:07:44 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:07:45 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:07:45 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 13:07:45 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:07:46 Main Volt Voltage Measurement, V = 0.000001 V 13:07:46 Temperature Measurement, Probe Number = 0, T = 29.477000 C 13:07:50 Temperature Measurement, Probe Number = 1, T = 29.238000 C 13:07:51 Temperature Measurement, Probe Number = 2, T = 28.625000 C 13:07:51 Temperature Measurement, Probe Number = 3, T = 28.831000 C 13:07:52 Temperature Measurement, Probe Number = 4, T = 28.926000 C 13:07:52 Temperature Measurement, Probe Number = 5, T = 26.110000 C 13:07:52 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.48 P3 = RTD Outside, Value = 29.24 P4 = Supply Water, Value = 28.62 P5 = Return Water, Value = 28.83 P6 = @ RTD Outside, Value = 28.93 P7 = Ambient, Value = 26.11 13:07:52 3 Axis Hall Probe Field Measurement, Bx = -0.000010, By = 0.000030, Bz = -0.000020 T 13:08:00 Move Stage, position = 0.188000 13:08:02 Stage Position Measurement, pos = 0.1882800 m 13:08:02 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:08:03 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:08:04 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:08:05 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:08:05 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 13:08:05 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:08:06 Main Volt Voltage Measurement, V = 0.000002 V 13:08:07 Temperature Measurement, Probe Number = 0, T = 29.480000 C 13:08:10 Temperature Measurement, Probe Number = 1, T = 29.241000 C 13:08:11 Temperature Measurement, Probe Number = 2, T = 28.652000 C 13:08:11 Temperature Measurement, Probe Number = 3, T = 28.853000 C 13:08:12 Temperature Measurement, Probe Number = 4, T = 28.933000 C 13:08:13 Temperature Measurement, Probe Number = 5, T = 26.160000 C 13:08:13 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.48 P3 = RTD Outside, Value = 29.24 P4 = Supply Water, Value = 28.65 P5 = Return Water, Value = 28.85 P6 = @ RTD Outside, Value = 28.93 P7 = Ambient, Value = 26.16 13:08:13 3 Axis Hall Probe Field Measurement, Bx = -0.000010, By = 0.000030, Bz = -0.000020 T 13:08:20 Move Stage, position = 0.190000 13:08:22 Stage Position Measurement, pos = 0.1902650 m 13:08:23 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:08:23 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:08:24 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:08:25 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:08:25 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 13:08:26 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:08:26 Main Volt Voltage Measurement, V = 0.000001 V 13:08:27 Temperature Measurement, Probe Number = 0, T = 29.482000 C 13:08:30 Temperature Measurement, Probe Number = 1, T = 29.243000 C 13:08:31 Temperature Measurement, Probe Number = 2, T = 28.656000 C 13:08:32 Temperature Measurement, Probe Number = 3, T = 28.872000 C 13:08:32 Temperature Measurement, Probe Number = 4, T = 28.933000 C 13:08:33 Temperature Measurement, Probe Number = 5, T = 26.141000 C 13:08:33 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.48 P3 = RTD Outside, Value = 29.24 P4 = Supply Water, Value = 28.66 P5 = Return Water, Value = 28.87 P6 = @ RTD Outside, Value = 28.93 P7 = Ambient, Value = 26.14 13:08:33 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000030, Bz = -0.000020 T 13:08:40 Move Stage, position = 0.192000 13:08:42 Stage Position Measurement, pos = 0.1922500 m 13:08:43 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:08:44 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:08:44 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:08:45 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:08:45 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 13:08:46 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:08:46 Main Volt Voltage Measurement, V = 0.000002 V 13:08:47 Temperature Measurement, Probe Number = 0, T = 29.484000 C 13:08:51 Temperature Measurement, Probe Number = 1, T = 29.245000 C 13:08:51 Temperature Measurement, Probe Number = 2, T = 28.666000 C 13:08:52 Temperature Measurement, Probe Number = 3, T = 28.877000 C 13:08:52 Temperature Measurement, Probe Number = 4, T = 28.940000 C 13:08:53 Temperature Measurement, Probe Number = 5, T = 26.170000 C 13:08:53 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.48 P3 = RTD Outside, Value = 29.25 P4 = Supply Water, Value = 28.67 P5 = Return Water, Value = 28.88 P6 = @ RTD Outside, Value = 28.94 P7 = Ambient, Value = 26.17 13:08:53 3 Axis Hall Probe Field Measurement, Bx = -0.000008, By = 0.000040, Bz = -0.000020 T 13:09:01 Move Stage, position = 0.194000 13:09:02 Stage Position Measurement, pos = 0.1942500 m 13:09:03 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:09:04 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:09:04 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:09:05 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:09:05 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 13:09:06 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:09:06 Main Volt Voltage Measurement, V = 0.000001 V 13:09:07 Temperature Measurement, Probe Number = 0, T = 29.486000 C 13:09:11 Temperature Measurement, Probe Number = 1, T = 29.239000 C 13:09:11 Temperature Measurement, Probe Number = 2, T = 28.694000 C 13:09:12 Temperature Measurement, Probe Number = 3, T = 28.884000 C 13:09:12 Temperature Measurement, Probe Number = 4, T = 28.936000 C 13:09:13 Temperature Measurement, Probe Number = 5, T = 26.276000 C 13:09:13 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.49 P3 = RTD Outside, Value = 29.24 P4 = Supply Water, Value = 28.69 P5 = Return Water, Value = 28.88 P6 = @ RTD Outside, Value = 28.94 P7 = Ambient, Value = 26.28 13:09:13 3 Axis Hall Probe Field Measurement, Bx = -0.000020, By = 0.000040, Bz = -0.000010 T 13:09:21 Move Stage, position = 0.196000 13:09:22 Stage Position Measurement, pos = 0.1962600 m 13:09:23 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:09:24 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:09:25 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:09:25 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:09:25 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 13:09:26 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:09:27 Main Volt Voltage Measurement, V = 0.000001 V 13:09:27 Temperature Measurement, Probe Number = 0, T = 29.486000 C 13:09:31 Temperature Measurement, Probe Number = 1, T = 29.227000 C 13:09:31 Temperature Measurement, Probe Number = 2, T = 28.697000 C 13:09:32 Temperature Measurement, Probe Number = 3, T = 28.898000 C 13:09:32 Temperature Measurement, Probe Number = 4, T = 28.916000 C 13:09:33 Temperature Measurement, Probe Number = 5, T = 26.291000 C 13:09:33 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.49 P3 = RTD Outside, Value = 29.23 P4 = Supply Water, Value = 28.70 P5 = Return Water, Value = 28.90 P6 = @ RTD Outside, Value = 28.92 P7 = Ambient, Value = 26.29 13:09:33 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000030, Bz = -0.000020 T 13:09:41 Move Stage, position = 0.198000 13:09:42 Stage Position Measurement, pos = 0.1982500 m 13:09:43 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:09:44 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:09:45 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:09:45 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:09:45 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 13:09:46 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:09:47 Main Volt Voltage Measurement, V = 0.000001 V 13:09:47 Temperature Measurement, Probe Number = 0, T = 29.487000 C 13:09:51 Temperature Measurement, Probe Number = 1, T = 29.231000 C 13:09:51 Temperature Measurement, Probe Number = 2, T = 28.673000 C 13:09:52 Temperature Measurement, Probe Number = 3, T = 28.904000 C 13:09:53 Temperature Measurement, Probe Number = 4, T = 28.908000 C 13:09:53 Temperature Measurement, Probe Number = 5, T = 26.225000 C 13:09:53 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.49 P3 = RTD Outside, Value = 29.23 P4 = Supply Water, Value = 28.67 P5 = Return Water, Value = 28.90 P6 = @ RTD Outside, Value = 28.91 P7 = Ambient, Value = 26.23 13:09:53 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000040, Bz = -0.000020 T 13:10:01 Move Stage, position = 0.200000 13:10:03 Stage Position Measurement, pos = 0.2002700 m 13:10:03 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:10:04 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:10:05 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:10:05 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:10:05 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 13:10:06 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:10:07 Main Volt Voltage Measurement, V = 0.000001 V 13:10:07 Temperature Measurement, Probe Number = 0, T = 29.490000 C 13:10:11 Temperature Measurement, Probe Number = 1, T = 29.248000 C 13:10:12 Temperature Measurement, Probe Number = 2, T = 28.662000 C 13:10:12 Temperature Measurement, Probe Number = 3, T = 28.903000 C 13:10:13 Temperature Measurement, Probe Number = 4, T = 28.921000 C 13:10:13 Temperature Measurement, Probe Number = 5, T = 26.214000 C 13:10:13 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.49 P3 = RTD Outside, Value = 29.25 P4 = Supply Water, Value = 28.66 P5 = Return Water, Value = 28.90 P6 = @ RTD Outside, Value = 28.92 P7 = Ambient, Value = 26.21 13:10:13 3 Axis Hall Probe Field Measurement, Bx = -0.000010, By = 0.000040, Bz = -0.000030 T 13:10:21 Move Stage, position = 0.202000 13:10:23 Stage Position Measurement, pos = 0.2022750 m 13:10:23 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:10:24 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:10:25 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:10:26 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:10:26 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 13:10:26 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:10:27 Main Volt Voltage Measurement, V = 0.000002 V 13:10:27 Temperature Measurement, Probe Number = 0, T = 29.491000 C 13:10:31 Temperature Measurement, Probe Number = 1, T = 29.253000 C 13:10:32 Temperature Measurement, Probe Number = 2, T = 28.648000 C 13:10:32 Temperature Measurement, Probe Number = 3, T = 28.888000 C 13:10:33 Temperature Measurement, Probe Number = 4, T = 28.935000 C 13:10:33 Temperature Measurement, Probe Number = 5, T = 26.394000 C 13:10:33 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.49 P3 = RTD Outside, Value = 29.25 P4 = Supply Water, Value = 28.65 P5 = Return Water, Value = 28.89 P6 = @ RTD Outside, Value = 28.93 P7 = Ambient, Value = 26.39 13:10:33 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000030, Bz = -0.000020 T 13:10:41 Move Stage, position = 0.204000 13:10:43 Stage Position Measurement, pos = 0.2042600 m 13:10:43 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:10:44 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:10:45 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:10:46 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:10:46 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 13:10:46 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:10:47 Main Volt Voltage Measurement, V = 0.000001 V 13:10:48 Temperature Measurement, Probe Number = 0, T = 29.488000 C 13:10:51 Temperature Measurement, Probe Number = 1, T = 29.245000 C 13:10:52 Temperature Measurement, Probe Number = 2, T = 28.627000 C 13:10:52 Temperature Measurement, Probe Number = 3, T = 28.880000 C 13:10:53 Temperature Measurement, Probe Number = 4, T = 28.939000 C 13:10:53 Temperature Measurement, Probe Number = 5, T = 26.255000 C 13:10:53 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.49 P3 = RTD Outside, Value = 29.25 P4 = Supply Water, Value = 28.63 P5 = Return Water, Value = 28.88 P6 = @ RTD Outside, Value = 28.94 P7 = Ambient, Value = 26.25 13:10:54 3 Axis Hall Probe Field Measurement, Bx = -0.000010, By = 0.000040, Bz = -0.000020 T 13:11:01 Move Stage, position = 0.206000 13:11:03 Stage Position Measurement, pos = 0.2063100 m 13:11:04 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:11:04 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:11:05 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:11:06 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:11:06 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 13:11:06 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:11:07 Main Volt Voltage Measurement, V = 0.000001 V 13:11:08 Temperature Measurement, Probe Number = 0, T = 29.484000 C 13:11:11 Temperature Measurement, Probe Number = 1, T = 29.247000 C 13:11:12 Temperature Measurement, Probe Number = 2, T = 28.615000 C 13:11:12 Temperature Measurement, Probe Number = 3, T = 28.878000 C 13:11:13 Temperature Measurement, Probe Number = 4, T = 28.944000 C 13:11:14 Temperature Measurement, Probe Number = 5, T = 26.200000 C 13:11:14 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.48 P3 = RTD Outside, Value = 29.25 P4 = Supply Water, Value = 28.61 P5 = Return Water, Value = 28.88 P6 = @ RTD Outside, Value = 28.94 P7 = Ambient, Value = 26.20 13:11:14 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000030, Bz = -0.000030 T 13:11:21 Move Stage, position = 0.208000 13:11:23 Stage Position Measurement, pos = 0.2082850 m 13:11:24 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:11:25 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:11:25 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:11:26 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:11:26 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 13:11:27 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:11:27 Main Volt Voltage Measurement, V = 0.000002 V 13:11:28 Temperature Measurement, Probe Number = 0, T = 29.483000 C 13:11:31 Temperature Measurement, Probe Number = 1, T = 29.242000 C 13:11:32 Temperature Measurement, Probe Number = 2, T = 28.620000 C 13:11:33 Temperature Measurement, Probe Number = 3, T = 28.868000 C 13:11:33 Temperature Measurement, Probe Number = 4, T = 28.940000 C 13:11:34 Temperature Measurement, Probe Number = 5, T = 26.166000 C 13:11:34 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.48 P3 = RTD Outside, Value = 29.24 P4 = Supply Water, Value = 28.62 P5 = Return Water, Value = 28.87 P6 = @ RTD Outside, Value = 28.94 P7 = Ambient, Value = 26.17 13:11:34 3 Axis Hall Probe Field Measurement, Bx = -0.000010, By = 0.000020, Bz = -0.000030 T 13:11:42 Move Stage, position = 0.210000 13:11:43 Stage Position Measurement, pos = 0.2102750 m 13:11:44 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:11:45 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:11:45 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:11:46 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:11:46 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 13:11:47 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:11:47 Main Volt Voltage Measurement, V = 0.000002 V 13:11:48 Temperature Measurement, Probe Number = 0, T = 29.486000 C 13:11:52 Temperature Measurement, Probe Number = 1, T = 29.254000 C 13:11:52 Temperature Measurement, Probe Number = 2, T = 28.621000 C 13:11:53 Temperature Measurement, Probe Number = 3, T = 28.865000 C 13:11:53 Temperature Measurement, Probe Number = 4, T = 28.949000 C 13:11:54 Temperature Measurement, Probe Number = 5, T = 26.166000 C 13:11:54 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.49 P3 = RTD Outside, Value = 29.25 P4 = Supply Water, Value = 28.62 P5 = Return Water, Value = 28.86 P6 = @ RTD Outside, Value = 28.95 P7 = Ambient, Value = 26.17 13:11:54 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000040, Bz = -0.000020 T 13:12:02 Move Stage, position = 0.212000 13:12:03 Stage Position Measurement, pos = 0.2123100 m 13:12:04 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:12:05 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:12:05 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:12:06 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:12:06 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 13:12:07 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:12:07 Main Volt Voltage Measurement, V = 0.000002 V 13:12:08 Temperature Measurement, Probe Number = 0, T = 29.483000 C 13:12:12 Temperature Measurement, Probe Number = 1, T = 29.254000 C 13:12:12 Temperature Measurement, Probe Number = 2, T = 28.609000 C 13:12:13 Temperature Measurement, Probe Number = 3, T = 28.865000 C 13:12:13 Temperature Measurement, Probe Number = 4, T = 28.957000 C 13:12:14 Temperature Measurement, Probe Number = 5, T = 26.276000 C 13:12:14 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.48 P3 = RTD Outside, Value = 29.25 P4 = Supply Water, Value = 28.61 P5 = Return Water, Value = 28.86 P6 = @ RTD Outside, Value = 28.96 P7 = Ambient, Value = 26.28 13:12:14 3 Axis Hall Probe Field Measurement, Bx = -0.000010, By = 0.000030, Bz = -0.000020 T 13:12:22 Move Stage, position = 0.214000 13:12:24 Stage Position Measurement, pos = 0.2143250 m 13:12:24 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:12:25 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:12:26 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:12:26 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:12:26 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 13:12:27 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:12:28 Main Volt Voltage Measurement, V = 0.000001 V 13:12:28 Temperature Measurement, Probe Number = 0, T = 29.480000 C 13:12:32 Temperature Measurement, Probe Number = 1, T = 29.252000 C 13:12:33 Temperature Measurement, Probe Number = 2, T = 28.622000 C 13:12:33 Temperature Measurement, Probe Number = 3, T = 28.864000 C 13:12:34 Temperature Measurement, Probe Number = 4, T = 28.959000 C 13:12:34 Temperature Measurement, Probe Number = 5, T = 26.279000 C 13:12:34 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.48 P3 = RTD Outside, Value = 29.25 P4 = Supply Water, Value = 28.62 P5 = Return Water, Value = 28.86 P6 = @ RTD Outside, Value = 28.96 P7 = Ambient, Value = 26.28 13:12:34 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000040, Bz = -0.000020 T 13:12:42 Move Stage, position = 0.216000 13:12:44 Stage Position Measurement, pos = 0.2163300 m 13:12:44 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:12:45 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:12:46 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:12:47 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:12:47 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 13:12:47 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:12:48 Main Volt Voltage Measurement, V = 0.000001 V 13:12:48 Temperature Measurement, Probe Number = 0, T = 29.478000 C 13:12:52 Temperature Measurement, Probe Number = 1, T = 29.244000 C 13:12:53 Temperature Measurement, Probe Number = 2, T = 28.634000 C 13:12:53 Temperature Measurement, Probe Number = 3, T = 28.870000 C 13:12:54 Temperature Measurement, Probe Number = 4, T = 28.956000 C 13:12:54 Temperature Measurement, Probe Number = 5, T = 26.245000 C 13:12:54 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.48 P3 = RTD Outside, Value = 29.24 P4 = Supply Water, Value = 28.63 P5 = Return Water, Value = 28.87 P6 = @ RTD Outside, Value = 28.96 P7 = Ambient, Value = 26.25 13:12:54 3 Axis Hall Probe Field Measurement, Bx = -0.000010, By = 0.000040, Bz = -0.000020 T 13:13:02 Move Stage, position = 0.218000 13:13:04 Stage Position Measurement, pos = 0.2183450 m 13:13:04 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:13:05 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:13:06 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:13:07 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:13:07 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 13:13:07 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:13:08 Main Volt Voltage Measurement, V = 0.000002 V 13:13:09 Temperature Measurement, Probe Number = 0, T = 29.480000 C 13:13:12 Temperature Measurement, Probe Number = 1, T = 29.244000 C 13:13:13 Temperature Measurement, Probe Number = 2, T = 28.639000 C 13:13:13 Temperature Measurement, Probe Number = 3, T = 28.866000 C 13:13:14 Temperature Measurement, Probe Number = 4, T = 28.951000 C 13:13:14 Temperature Measurement, Probe Number = 5, T = 26.199000 C 13:13:14 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.48 P3 = RTD Outside, Value = 29.24 P4 = Supply Water, Value = 28.64 P5 = Return Water, Value = 28.87 P6 = @ RTD Outside, Value = 28.95 P7 = Ambient, Value = 26.20 13:13:15 3 Axis Hall Probe Field Measurement, Bx = -0.000010, By = 0.000040, Bz = -0.000020 T 13:13:22 Move Stage, position = 0.220000 13:13:24 Stage Position Measurement, pos = 0.2203550 m 13:13:25 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:13:25 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:13:26 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:13:27 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:13:27 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 13:13:27 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:13:28 Main Volt Voltage Measurement, V = 0.000002 V 13:13:29 Temperature Measurement, Probe Number = 0, T = 29.483000 C 13:13:32 Temperature Measurement, Probe Number = 1, T = 29.246000 C 13:13:33 Temperature Measurement, Probe Number = 2, T = 28.640000 C 13:13:33 Temperature Measurement, Probe Number = 3, T = 28.851000 C 13:13:34 Temperature Measurement, Probe Number = 4, T = 28.953000 C 13:13:35 Temperature Measurement, Probe Number = 5, T = 26.233000 C 13:13:35 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.48 P3 = RTD Outside, Value = 29.25 P4 = Supply Water, Value = 28.64 P5 = Return Water, Value = 28.85 P6 = @ RTD Outside, Value = 28.95 P7 = Ambient, Value = 26.23 13:13:35 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000040, Bz = -0.000020 T 13:13:42 Move Stage, position = 0.222000 13:13:44 Stage Position Measurement, pos = 0.2223650 m 13:13:45 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:13:45 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:13:46 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:13:47 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:13:47 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 13:13:48 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:13:48 Main Volt Voltage Measurement, V = 0.000002 V 13:13:49 Temperature Measurement, Probe Number = 0, T = 29.486000 C 13:13:52 Temperature Measurement, Probe Number = 1, T = 29.251000 C 13:13:53 Temperature Measurement, Probe Number = 2, T = 28.630000 C 13:13:54 Temperature Measurement, Probe Number = 3, T = 28.845000 C 13:13:54 Temperature Measurement, Probe Number = 4, T = 28.958000 C 13:13:55 Temperature Measurement, Probe Number = 5, T = 26.248000 C 13:13:55 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.49 P3 = RTD Outside, Value = 29.25 P4 = Supply Water, Value = 28.63 P5 = Return Water, Value = 28.84 P6 = @ RTD Outside, Value = 28.96 P7 = Ambient, Value = 26.25 13:13:55 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000030, Bz = -0.000020 T 13:14:02 Move Stage, position = 0.224000 13:14:04 Stage Position Measurement, pos = 0.2243850 m 13:14:05 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:14:06 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:14:06 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:14:07 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:14:07 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 13:14:08 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:14:08 Main Volt Voltage Measurement, V = 0.000001 V 13:14:09 Temperature Measurement, Probe Number = 0, T = 29.486000 C 13:14:12 Temperature Measurement, Probe Number = 1, T = 29.257000 C 13:14:13 Temperature Measurement, Probe Number = 2, T = 28.592000 C 13:14:14 Temperature Measurement, Probe Number = 3, T = 28.828000 C 13:14:14 Temperature Measurement, Probe Number = 4, T = 28.968000 C 13:14:15 Temperature Measurement, Probe Number = 5, T = 26.142000 C 13:14:15 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.49 P3 = RTD Outside, Value = 29.26 P4 = Supply Water, Value = 28.59 P5 = Return Water, Value = 28.83 P6 = @ RTD Outside, Value = 28.97 P7 = Ambient, Value = 26.14 13:14:15 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000040, Bz = -0.000030 T 13:14:23 Move Stage, position = 0.226000 13:14:24 Stage Position Measurement, pos = 0.2263800 m 13:14:25 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:14:26 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:14:26 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:14:27 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:14:27 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 13:14:28 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:14:28 Main Volt Voltage Measurement, V = 0.000001 V 13:14:29 Temperature Measurement, Probe Number = 0, T = 29.485000 C 13:14:33 Temperature Measurement, Probe Number = 1, T = 29.257000 C 13:14:33 Temperature Measurement, Probe Number = 2, T = 28.548000 C 13:14:34 Temperature Measurement, Probe Number = 3, T = 28.815000 C 13:14:34 Temperature Measurement, Probe Number = 4, T = 28.972000 C 13:14:35 Temperature Measurement, Probe Number = 5, T = 26.069000 C 13:14:35 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.48 P3 = RTD Outside, Value = 29.26 P4 = Supply Water, Value = 28.55 P5 = Return Water, Value = 28.82 P6 = @ RTD Outside, Value = 28.97 P7 = Ambient, Value = 26.07 13:14:35 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000040, Bz = -0.000030 T 13:14:43 Move Stage, position = 0.228000 13:14:44 Stage Position Measurement, pos = 0.2283650 m 13:14:45 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:14:46 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:14:46 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:14:47 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:14:47 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 13:14:48 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:14:48 Main Volt Voltage Measurement, V = 0.000002 V 13:14:49 Temperature Measurement, Probe Number = 0, T = 29.482000 C 13:14:53 Temperature Measurement, Probe Number = 1, T = 29.253000 C 13:14:53 Temperature Measurement, Probe Number = 2, T = 28.518000 C 13:14:54 Temperature Measurement, Probe Number = 3, T = 28.822000 C 13:14:54 Temperature Measurement, Probe Number = 4, T = 28.972000 C 13:14:55 Temperature Measurement, Probe Number = 5, T = 26.095000 C 13:14:55 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.48 P3 = RTD Outside, Value = 29.25 P4 = Supply Water, Value = 28.52 P5 = Return Water, Value = 28.82 P6 = @ RTD Outside, Value = 28.97 P7 = Ambient, Value = 26.09 13:14:55 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000040, Bz = -0.000030 T 13:15:03 Move Stage, position = 0.230000 13:15:04 Stage Position Measurement, pos = 0.2303300 m 13:15:05 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:15:06 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:15:07 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:15:07 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:15:07 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 13:15:08 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:15:09 Main Volt Voltage Measurement, V = 0.000002 V 13:15:09 Temperature Measurement, Probe Number = 0, T = 29.477000 C 13:15:13 Temperature Measurement, Probe Number = 1, T = 29.244000 C 13:15:13 Temperature Measurement, Probe Number = 2, T = 28.515000 C 13:15:14 Temperature Measurement, Probe Number = 3, T = 28.828000 C 13:15:15 Temperature Measurement, Probe Number = 4, T = 28.966000 C 13:15:15 Temperature Measurement, Probe Number = 5, T = 26.143000 C 13:15:15 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.48 P3 = RTD Outside, Value = 29.24 P4 = Supply Water, Value = 28.52 P5 = Return Water, Value = 28.83 P6 = @ RTD Outside, Value = 28.97 P7 = Ambient, Value = 26.14 13:15:15 3 Axis Hall Probe Field Measurement, Bx = -0.000010, By = 0.000040, Bz = -0.000020 T 13:15:23 Move Stage, position = 0.232000 13:15:24 Stage Position Measurement, pos = 0.2323050 m 13:15:25 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:15:26 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:15:27 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:15:27 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:15:27 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 13:15:28 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:15:29 Main Volt Voltage Measurement, V = 0.000002 V 13:15:29 Temperature Measurement, Probe Number = 0, T = 29.477000 C 13:15:33 Temperature Measurement, Probe Number = 1, T = 29.237000 C 13:15:34 Temperature Measurement, Probe Number = 2, T = 28.537000 C 13:15:34 Temperature Measurement, Probe Number = 3, T = 28.848000 C 13:15:35 Temperature Measurement, Probe Number = 4, T = 28.955000 C 13:15:35 Temperature Measurement, Probe Number = 5, T = 26.202000 C 13:15:35 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.48 P3 = RTD Outside, Value = 29.24 P4 = Supply Water, Value = 28.54 P5 = Return Water, Value = 28.85 P6 = @ RTD Outside, Value = 28.95 P7 = Ambient, Value = 26.20 13:15:35 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000040, Bz = -0.000030 T 13:15:43 Move Stage, position = 0.234000 13:15:45 Stage Position Measurement, pos = 0.2342800 m 13:15:45 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:15:46 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:15:47 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:15:48 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:15:48 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 13:15:48 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:15:49 Main Volt Voltage Measurement, V = 0.000001 V 13:15:49 Temperature Measurement, Probe Number = 0, T = 29.481000 C 13:15:53 Temperature Measurement, Probe Number = 1, T = 29.249000 C 13:15:54 Temperature Measurement, Probe Number = 2, T = 28.551000 C 13:15:54 Temperature Measurement, Probe Number = 3, T = 28.862000 C 13:15:55 Temperature Measurement, Probe Number = 4, T = 28.957000 C 13:15:55 Temperature Measurement, Probe Number = 5, T = 26.189000 C 13:15:55 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.48 P3 = RTD Outside, Value = 29.25 P4 = Supply Water, Value = 28.55 P5 = Return Water, Value = 28.86 P6 = @ RTD Outside, Value = 28.96 P7 = Ambient, Value = 26.19 13:15:55 3 Axis Hall Probe Field Measurement, Bx = 0.000010, By = 0.000050, Bz = -0.000020 T 13:16:03 Move Stage, position = 0.236000 13:16:05 Stage Position Measurement, pos = 0.2362900 m 13:16:06 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:16:06 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:16:07 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:16:08 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:16:08 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 13:16:08 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:16:09 Main Volt Voltage Measurement, V = 0.000001 V 13:16:10 Temperature Measurement, Probe Number = 0, T = 29.482000 C 13:16:13 Temperature Measurement, Probe Number = 1, T = 29.251000 C 13:16:14 Temperature Measurement, Probe Number = 2, T = 28.537000 C 13:16:14 Temperature Measurement, Probe Number = 3, T = 28.864000 C 13:16:15 Temperature Measurement, Probe Number = 4, T = 28.960000 C 13:16:16 Temperature Measurement, Probe Number = 5, T = 26.067000 C 13:16:16 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.48 P3 = RTD Outside, Value = 29.25 P4 = Supply Water, Value = 28.54 P5 = Return Water, Value = 28.86 P6 = @ RTD Outside, Value = 28.96 P7 = Ambient, Value = 26.07 13:16:16 3 Axis Hall Probe Field Measurement, Bx = -0.000010, By = 0.000040, Bz = -0.000020 T 13:16:23 Move Stage, position = 0.238000 13:16:25 Stage Position Measurement, pos = 0.2383000 m 13:16:26 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:16:26 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:16:27 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:16:28 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:16:28 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 13:16:28 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:16:29 Main Volt Voltage Measurement, V = 0.000002 V 13:16:30 Temperature Measurement, Probe Number = 0, T = 29.484000 C 13:16:33 Temperature Measurement, Probe Number = 1, T = 29.253000 C 13:16:34 Temperature Measurement, Probe Number = 2, T = 28.509000 C 13:16:34 Temperature Measurement, Probe Number = 3, T = 28.841000 C 13:16:35 Temperature Measurement, Probe Number = 4, T = 28.965000 C 13:16:36 Temperature Measurement, Probe Number = 5, T = 26.107000 C 13:16:36 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.48 P3 = RTD Outside, Value = 29.25 P4 = Supply Water, Value = 28.51 P5 = Return Water, Value = 28.84 P6 = @ RTD Outside, Value = 28.96 P7 = Ambient, Value = 26.11 13:16:36 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000040, Bz = -0.000010 T 13:16:43 Move Stage, position = 0.240000 13:16:45 Stage Position Measurement, pos = 0.2403050 m 13:16:46 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:16:46 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:16:47 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:16:48 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:16:48 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 13:16:49 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:16:49 Main Volt Voltage Measurement, V = 0.000002 V 13:16:50 Temperature Measurement, Probe Number = 0, T = 29.487000 C 13:16:53 Temperature Measurement, Probe Number = 1, T = 29.259000 C 13:16:54 Temperature Measurement, Probe Number = 2, T = 28.511000 C 13:16:55 Temperature Measurement, Probe Number = 3, T = 28.836000 C 13:16:55 Temperature Measurement, Probe Number = 4, T = 28.974000 C 13:16:56 Temperature Measurement, Probe Number = 5, T = 26.131000 C 13:16:56 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.49 P3 = RTD Outside, Value = 29.26 P4 = Supply Water, Value = 28.51 P5 = Return Water, Value = 28.84 P6 = @ RTD Outside, Value = 28.97 P7 = Ambient, Value = 26.13 13:16:56 3 Axis Hall Probe Field Measurement, Bx = -0.000010, By = 0.000030, Bz = -0.000030 T 13:17:04 Move Stage, position = 0.242000 13:17:05 Stage Position Measurement, pos = 0.2423250 m 13:17:06 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:17:07 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:17:07 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:17:08 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:17:08 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 13:17:09 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:17:09 Main Volt Voltage Measurement, V = 0.000002 V 13:17:10 Temperature Measurement, Probe Number = 0, T = 29.490000 C 13:17:14 Temperature Measurement, Probe Number = 1, T = 29.258000 C 13:17:14 Temperature Measurement, Probe Number = 2, T = 28.537000 C 13:17:15 Temperature Measurement, Probe Number = 3, T = 28.836000 C 13:17:15 Temperature Measurement, Probe Number = 4, T = 28.981000 C 13:17:16 Temperature Measurement, Probe Number = 5, T = 26.154000 C 13:17:16 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.49 P3 = RTD Outside, Value = 29.26 P4 = Supply Water, Value = 28.54 P5 = Return Water, Value = 28.84 P6 = @ RTD Outside, Value = 28.98 P7 = Ambient, Value = 26.15 13:17:16 3 Axis Hall Probe Field Measurement, Bx = -0.000010, By = 0.000040, Bz = -0.000030 T 13:17:24 Move Stage, position = 0.244000 13:17:25 Stage Position Measurement, pos = 0.2443000 m 13:17:26 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:17:27 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:17:27 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:17:28 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:17:28 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 13:17:29 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:17:29 Main Volt Voltage Measurement, V = 0.000002 V 13:17:30 Temperature Measurement, Probe Number = 0, T = 29.487000 C 13:17:34 Temperature Measurement, Probe Number = 1, T = 29.252000 C 13:17:34 Temperature Measurement, Probe Number = 2, T = 28.571000 C 13:17:35 Temperature Measurement, Probe Number = 3, T = 28.850000 C 13:17:35 Temperature Measurement, Probe Number = 4, T = 28.981000 C 13:17:36 Temperature Measurement, Probe Number = 5, T = 26.241000 C 13:17:36 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.49 P3 = RTD Outside, Value = 29.25 P4 = Supply Water, Value = 28.57 P5 = Return Water, Value = 28.85 P6 = @ RTD Outside, Value = 28.98 P7 = Ambient, Value = 26.24 13:17:36 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000040, Bz = -0.000020 T 13:17:44 Move Stage, position = 0.246000 13:17:45 Stage Position Measurement, pos = 0.2463050 m 13:17:46 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:17:47 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:17:47 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:17:48 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:17:48 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 13:17:49 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:17:49 Main Volt Voltage Measurement, V = 0.000001 V 13:17:50 Temperature Measurement, Probe Number = 0, T = 29.486000 C 13:17:54 Temperature Measurement, Probe Number = 1, T = 29.248000 C 13:17:54 Temperature Measurement, Probe Number = 2, T = 28.592000 C 13:17:55 Temperature Measurement, Probe Number = 3, T = 28.861000 C 13:17:55 Temperature Measurement, Probe Number = 4, T = 28.975000 C 13:17:56 Temperature Measurement, Probe Number = 5, T = 26.249000 C 13:17:56 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.49 P3 = RTD Outside, Value = 29.25 P4 = Supply Water, Value = 28.59 P5 = Return Water, Value = 28.86 P6 = @ RTD Outside, Value = 28.98 P7 = Ambient, Value = 26.25 13:17:56 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000040, Bz = -0.000030 T 13:18:04 Move Stage, position = 0.248000 13:18:05 Stage Position Measurement, pos = 0.2482900 m 13:18:06 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:18:07 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:18:08 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:18:08 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:18:08 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 13:18:09 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:18:10 Main Volt Voltage Measurement, V = 0.000001 V 13:18:10 Temperature Measurement, Probe Number = 0, T = 29.487000 C 13:18:14 Temperature Measurement, Probe Number = 1, T = 29.234000 C 13:18:14 Temperature Measurement, Probe Number = 2, T = 28.613000 C 13:18:15 Temperature Measurement, Probe Number = 3, T = 28.874000 C 13:18:16 Temperature Measurement, Probe Number = 4, T = 28.965000 C 13:18:16 Temperature Measurement, Probe Number = 5, T = 26.183000 C 13:18:16 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.49 P3 = RTD Outside, Value = 29.23 P4 = Supply Water, Value = 28.61 P5 = Return Water, Value = 28.87 P6 = @ RTD Outside, Value = 28.96 P7 = Ambient, Value = 26.18 13:18:16 3 Axis Hall Probe Field Measurement, Bx = -0.000010, By = 0.000030, Bz = -0.000020 T 13:18:24 Move Stage, position = 0.250000 13:18:26 Stage Position Measurement, pos = 0.2502850 m 13:18:26 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:18:27 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:18:28 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:18:28 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:18:28 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 13:18:29 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:18:30 Main Volt Voltage Measurement, V = 0.000002 V 13:18:30 Temperature Measurement, Probe Number = 0, T = 29.485000 C 13:18:34 Temperature Measurement, Probe Number = 1, T = 29.236000 C 13:18:34 Temperature Measurement, Probe Number = 2, T = 28.614000 C 13:18:35 Temperature Measurement, Probe Number = 3, T = 28.857000 C 13:18:36 Temperature Measurement, Probe Number = 4, T = 28.958000 C 13:18:36 Temperature Measurement, Probe Number = 5, T = 26.160000 C 13:18:36 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.48 P3 = RTD Outside, Value = 29.24 P4 = Supply Water, Value = 28.61 P5 = Return Water, Value = 28.86 P6 = @ RTD Outside, Value = 28.96 P7 = Ambient, Value = 26.16 13:18:36 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000030, Bz = -0.000020 T 13:18:44 Move Stage, position = 0.252000 13:18:46 Stage Position Measurement, pos = 0.2522800 m 13:18:46 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:18:47 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:18:48 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:18:48 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:18:48 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 13:18:49 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:18:50 Main Volt Voltage Measurement, V = 0.000002 V 13:18:50 Temperature Measurement, Probe Number = 0, T = 29.484000 C 13:18:54 Temperature Measurement, Probe Number = 1, T = 29.244000 C 13:18:55 Temperature Measurement, Probe Number = 2, T = 28.644000 C 13:18:55 Temperature Measurement, Probe Number = 3, T = 28.825000 C 13:18:56 Temperature Measurement, Probe Number = 4, T = 28.949000 C 13:18:56 Temperature Measurement, Probe Number = 5, T = 26.166000 C 13:18:56 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.48 P3 = RTD Outside, Value = 29.24 P4 = Supply Water, Value = 28.64 P5 = Return Water, Value = 28.82 P6 = @ RTD Outside, Value = 28.95 P7 = Ambient, Value = 26.17 13:18:56 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000040, Bz = -0.000020 T 13:19:04 Move Stage, position = 0.254000 13:19:06 Stage Position Measurement, pos = 0.2542800 m 13:19:06 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:19:07 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:19:08 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:19:09 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:19:09 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 13:19:09 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:19:10 Main Volt Voltage Measurement, V = 0.000002 V 13:19:11 Temperature Measurement, Probe Number = 0, T = 29.482000 C 13:19:14 Temperature Measurement, Probe Number = 1, T = 29.250000 C 13:19:15 Temperature Measurement, Probe Number = 2, T = 28.680000 C 13:19:15 Temperature Measurement, Probe Number = 3, T = 28.796000 C 13:19:16 Temperature Measurement, Probe Number = 4, T = 28.957000 C 13:19:16 Temperature Measurement, Probe Number = 5, T = 26.150000 C 13:19:16 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.48 P3 = RTD Outside, Value = 29.25 P4 = Supply Water, Value = 28.68 P5 = Return Water, Value = 28.80 P6 = @ RTD Outside, Value = 28.96 P7 = Ambient, Value = 26.15 13:19:17 3 Axis Hall Probe Field Measurement, Bx = -0.000010, By = 0.000030, Bz = -0.000030 T 13:19:24 Move Stage, position = 0.256000 13:19:26 Stage Position Measurement, pos = 0.2563200 m 13:19:27 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:19:27 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:19:28 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:19:29 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:19:29 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 13:19:29 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:19:30 Main Volt Voltage Measurement, V = 0.000002 V 13:19:31 Temperature Measurement, Probe Number = 0, T = 29.480000 C 13:19:34 Temperature Measurement, Probe Number = 1, T = 29.251000 C 13:19:35 Temperature Measurement, Probe Number = 2, T = 28.705000 C 13:19:35 Temperature Measurement, Probe Number = 3, T = 28.790000 C 13:19:36 Temperature Measurement, Probe Number = 4, T = 28.964000 C 13:19:37 Temperature Measurement, Probe Number = 5, T = 26.152000 C 13:19:37 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.48 P3 = RTD Outside, Value = 29.25 P4 = Supply Water, Value = 28.70 P5 = Return Water, Value = 28.79 P6 = @ RTD Outside, Value = 28.96 P7 = Ambient, Value = 26.15 13:19:37 3 Axis Hall Probe Field Measurement, Bx = -0.000010, By = 0.000040, Bz = -0.000030 T 13:19:44 Move Stage, position = 0.258000 13:19:46 Stage Position Measurement, pos = 0.2583100 m 13:19:47 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:19:47 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:19:48 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:19:49 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:19:49 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 13:19:50 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:19:50 Main Volt Voltage Measurement, V = 0.000001 V 13:19:51 Temperature Measurement, Probe Number = 0, T = 29.480000 C 13:19:54 Temperature Measurement, Probe Number = 1, T = 29.246000 C 13:19:55 Temperature Measurement, Probe Number = 2, T = 28.732000 C 13:19:56 Temperature Measurement, Probe Number = 3, T = 28.765000 C 13:19:56 Temperature Measurement, Probe Number = 4, T = 28.964000 C 13:19:57 Temperature Measurement, Probe Number = 5, T = 26.157000 C 13:19:57 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.48 P3 = RTD Outside, Value = 29.25 P4 = Supply Water, Value = 28.73 P5 = Return Water, Value = 28.77 P6 = @ RTD Outside, Value = 28.96 P7 = Ambient, Value = 26.16 13:19:57 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000030, Bz = -0.000030 T 13:20:04 Move Stage, position = 0.260000 13:20:06 Stage Position Measurement, pos = 0.2603300 m 13:20:07 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:20:08 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:20:08 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:20:09 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:20:09 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 13:20:10 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:20:10 Main Volt Voltage Measurement, V = 0.000001 V 13:20:11 Temperature Measurement, Probe Number = 0, T = 29.482000 C 13:20:14 Temperature Measurement, Probe Number = 1, T = 29.241000 C 13:20:15 Temperature Measurement, Probe Number = 2, T = 28.756000 C 13:20:16 Temperature Measurement, Probe Number = 3, T = 28.769000 C 13:20:16 Temperature Measurement, Probe Number = 4, T = 28.956000 C 13:20:17 Temperature Measurement, Probe Number = 5, T = 26.180000 C 13:20:17 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.48 P3 = RTD Outside, Value = 29.24 P4 = Supply Water, Value = 28.76 P5 = Return Water, Value = 28.77 P6 = @ RTD Outside, Value = 28.96 P7 = Ambient, Value = 26.18 13:20:17 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000040, Bz = -0.000030 T 13:20:25 Move Stage, position = 0.262000 13:20:26 Stage Position Measurement, pos = 0.2623000 m 13:20:27 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:20:28 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:20:28 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:20:29 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:20:29 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 13:20:30 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:20:30 Main Volt Voltage Measurement, V = 0.000002 V 13:20:31 Temperature Measurement, Probe Number = 0, T = 29.480000 C 13:20:35 Temperature Measurement, Probe Number = 1, T = 29.218000 C 13:20:35 Temperature Measurement, Probe Number = 2, T = 28.763000 C 13:20:36 Temperature Measurement, Probe Number = 3, T = 28.793000 C 13:20:36 Temperature Measurement, Probe Number = 4, T = 28.937000 C 13:20:37 Temperature Measurement, Probe Number = 5, T = 26.191000 C 13:20:37 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.48 P3 = RTD Outside, Value = 29.22 P4 = Supply Water, Value = 28.76 P5 = Return Water, Value = 28.79 P6 = @ RTD Outside, Value = 28.94 P7 = Ambient, Value = 26.19 13:20:37 3 Axis Hall Probe Field Measurement, Bx = -0.000010, By = 0.000050, Bz = -0.000020 T 13:20:45 Move Stage, position = 0.264000 13:20:46 Stage Position Measurement, pos = 0.2643050 m 13:20:47 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:20:48 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:20:48 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:20:49 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:20:49 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 13:20:50 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:20:50 Main Volt Voltage Measurement, V = 0.000001 V 13:20:51 Temperature Measurement, Probe Number = 0, T = 29.481000 C 13:20:55 Temperature Measurement, Probe Number = 1, T = 29.232000 C 13:20:55 Temperature Measurement, Probe Number = 2, T = 28.754000 C 13:20:56 Temperature Measurement, Probe Number = 3, T = 28.809000 C 13:20:56 Temperature Measurement, Probe Number = 4, T = 28.936000 C 13:20:57 Temperature Measurement, Probe Number = 5, T = 26.207000 C 13:20:57 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.48 P3 = RTD Outside, Value = 29.23 P4 = Supply Water, Value = 28.75 P5 = Return Water, Value = 28.81 P6 = @ RTD Outside, Value = 28.94 P7 = Ambient, Value = 26.21 13:20:57 3 Axis Hall Probe Field Measurement, Bx = -0.000010, By = 0.000050, Bz = -0.000020 T 13:21:05 Move Stage, position = 0.266000 13:21:06 Stage Position Measurement, pos = 0.2662900 m 13:21:07 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:21:08 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:21:09 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:21:09 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:21:09 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 13:21:10 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:21:11 Main Volt Voltage Measurement, V = 0.000001 V 13:21:11 Temperature Measurement, Probe Number = 0, T = 29.482000 C 13:21:15 Temperature Measurement, Probe Number = 1, T = 29.244000 C 13:21:15 Temperature Measurement, Probe Number = 2, T = 28.770000 C 13:21:16 Temperature Measurement, Probe Number = 3, T = 28.824000 C 13:21:17 Temperature Measurement, Probe Number = 4, T = 28.946000 C 13:21:17 Temperature Measurement, Probe Number = 5, T = 26.243000 C 13:21:17 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.48 P3 = RTD Outside, Value = 29.24 P4 = Supply Water, Value = 28.77 P5 = Return Water, Value = 28.82 P6 = @ RTD Outside, Value = 28.95 P7 = Ambient, Value = 26.24 13:21:17 3 Axis Hall Probe Field Measurement, Bx = -0.000010, By = 0.000040, Bz = -0.000020 T 13:21:25 Move Stage, position = 0.268000 13:21:27 Stage Position Measurement, pos = 0.2682900 m 13:21:27 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:21:28 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:21:29 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:21:29 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:21:29 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 13:21:30 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:21:31 Main Volt Voltage Measurement, V = 0.000001 V 13:21:31 Temperature Measurement, Probe Number = 0, T = 29.481000 C 13:21:35 Temperature Measurement, Probe Number = 1, T = 29.250000 C 13:21:36 Temperature Measurement, Probe Number = 2, T = 28.793000 C 13:21:36 Temperature Measurement, Probe Number = 3, T = 28.835000 C 13:21:37 Temperature Measurement, Probe Number = 4, T = 28.957000 C 13:21:37 Temperature Measurement, Probe Number = 5, T = 26.281000 C 13:21:37 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.48 P3 = RTD Outside, Value = 29.25 P4 = Supply Water, Value = 28.79 P5 = Return Water, Value = 28.84 P6 = @ RTD Outside, Value = 28.96 P7 = Ambient, Value = 26.28 13:21:37 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000040, Bz = -0.000020 T 13:21:45 Move Stage, position = 0.270000 13:21:47 Stage Position Measurement, pos = 0.2702850 m 13:21:47 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:21:48 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:21:49 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:21:50 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:21:50 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 13:21:50 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:21:51 Main Volt Voltage Measurement, V = 0.000001 V 13:21:51 Temperature Measurement, Probe Number = 0, T = 29.480000 C 13:21:55 Temperature Measurement, Probe Number = 1, T = 29.251000 C 13:21:56 Temperature Measurement, Probe Number = 2, T = 28.811000 C 13:21:56 Temperature Measurement, Probe Number = 3, T = 28.849000 C 13:21:57 Temperature Measurement, Probe Number = 4, T = 28.968000 C 13:21:57 Temperature Measurement, Probe Number = 5, T = 26.252000 C 13:21:57 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.48 P3 = RTD Outside, Value = 29.25 P4 = Supply Water, Value = 28.81 P5 = Return Water, Value = 28.85 P6 = @ RTD Outside, Value = 28.97 P7 = Ambient, Value = 26.25 13:21:57 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000040, Bz = -0.000020 T 13:22:05 Move Stage, position = 0.272000 13:22:07 Stage Position Measurement, pos = 0.2722750 m 13:22:07 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:22:08 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:22:09 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:22:10 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:22:10 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 13:22:10 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:22:11 Main Volt Voltage Measurement, V = 0.000001 V 13:22:12 Temperature Measurement, Probe Number = 0, T = 29.481000 C 13:22:15 Temperature Measurement, Probe Number = 1, T = 29.254000 C 13:22:16 Temperature Measurement, Probe Number = 2, T = 28.782000 C 13:22:16 Temperature Measurement, Probe Number = 3, T = 28.845000 C 13:22:17 Temperature Measurement, Probe Number = 4, T = 28.977000 C 13:22:17 Temperature Measurement, Probe Number = 5, T = 26.222000 C 13:22:17 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.48 P3 = RTD Outside, Value = 29.25 P4 = Supply Water, Value = 28.78 P5 = Return Water, Value = 28.84 P6 = @ RTD Outside, Value = 28.98 P7 = Ambient, Value = 26.22 13:22:18 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000050, Bz = -0.000020 T 13:22:25 Move Stage, position = 0.274000 13:22:27 Stage Position Measurement, pos = 0.2743050 m 13:22:28 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:22:28 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:22:29 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:22:30 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:22:30 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 13:22:30 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:22:31 Main Volt Voltage Measurement, V = 0.000001 V 13:22:32 Temperature Measurement, Probe Number = 0, T = 29.481000 C 13:22:35 Temperature Measurement, Probe Number = 1, T = 29.253000 C 13:22:36 Temperature Measurement, Probe Number = 2, T = 28.759000 C 13:22:36 Temperature Measurement, Probe Number = 3, T = 28.841000 C 13:22:37 Temperature Measurement, Probe Number = 4, T = 28.983000 C 13:22:38 Temperature Measurement, Probe Number = 5, T = 26.372000 C 13:22:38 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.48 P3 = RTD Outside, Value = 29.25 P4 = Supply Water, Value = 28.76 P5 = Return Water, Value = 28.84 P6 = @ RTD Outside, Value = 28.98 P7 = Ambient, Value = 26.37 13:22:38 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000040, Bz = -0.000020 T 13:22:45 Move Stage, position = 0.276000 13:22:47 Stage Position Measurement, pos = 0.2763200 m 13:22:48 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:22:48 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:22:49 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:22:50 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:22:50 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 13:22:51 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:22:51 Main Volt Voltage Measurement, V = 0.000001 V 13:22:52 Temperature Measurement, Probe Number = 0, T = 29.483000 C 13:22:55 Temperature Measurement, Probe Number = 1, T = 29.252000 C 13:22:56 Temperature Measurement, Probe Number = 2, T = 28.771000 C 13:22:57 Temperature Measurement, Probe Number = 3, T = 28.863000 C 13:22:57 Temperature Measurement, Probe Number = 4, T = 28.987000 C 13:22:58 Temperature Measurement, Probe Number = 5, T = 26.369000 C 13:22:58 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.48 P3 = RTD Outside, Value = 29.25 P4 = Supply Water, Value = 28.77 P5 = Return Water, Value = 28.86 P6 = @ RTD Outside, Value = 28.99 P7 = Ambient, Value = 26.37 13:22:58 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000040, Bz = -0.000020 T 13:23:05 Move Stage, position = 0.278000 13:23:07 Stage Position Measurement, pos = 0.2783400 m 13:23:08 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:23:09 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:23:09 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:23:10 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:23:10 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 13:23:11 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:23:11 Main Volt Voltage Measurement, V = 0.000001 V 13:23:12 Temperature Measurement, Probe Number = 0, T = 29.485000 C 13:23:15 Temperature Measurement, Probe Number = 1, T = 29.255000 C 13:23:16 Temperature Measurement, Probe Number = 2, T = 28.794000 C 13:23:17 Temperature Measurement, Probe Number = 3, T = 28.882000 C 13:23:17 Temperature Measurement, Probe Number = 4, T = 28.990000 C 13:23:18 Temperature Measurement, Probe Number = 5, T = 26.357000 C 13:23:18 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.48 P3 = RTD Outside, Value = 29.25 P4 = Supply Water, Value = 28.79 P5 = Return Water, Value = 28.88 P6 = @ RTD Outside, Value = 28.99 P7 = Ambient, Value = 26.36 13:23:18 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000040, Bz = -0.000030 T 13:23:26 Move Stage, position = 0.280000 13:23:27 Stage Position Measurement, pos = 0.2803150 m 13:23:28 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:23:29 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:23:29 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:23:30 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:23:30 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 13:23:31 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:23:31 Main Volt Voltage Measurement, V = 0.000001 V 13:23:32 Temperature Measurement, Probe Number = 0, T = 29.488000 C 13:23:36 Temperature Measurement, Probe Number = 1, T = 29.258000 C 13:23:36 Temperature Measurement, Probe Number = 2, T = 28.809000 C 13:23:37 Temperature Measurement, Probe Number = 3, T = 28.896000 C 13:23:37 Temperature Measurement, Probe Number = 4, T = 28.997000 C 13:23:38 Temperature Measurement, Probe Number = 5, T = 26.298000 C 13:23:38 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.49 P3 = RTD Outside, Value = 29.26 P4 = Supply Water, Value = 28.81 P5 = Return Water, Value = 28.90 P6 = @ RTD Outside, Value = 29.00 P7 = Ambient, Value = 26.30 13:23:38 3 Axis Hall Probe Field Measurement, Bx = -0.000010, By = 0.000040, Bz = -0.000030 T 13:23:46 Move Stage, position = 0.282000 13:23:47 Stage Position Measurement, pos = 0.2823000 m 13:23:48 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:23:49 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:23:49 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:23:50 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:23:50 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 13:23:51 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:23:51 Main Volt Voltage Measurement, V = 0.000001 V 13:23:52 Temperature Measurement, Probe Number = 0, T = 29.488000 C 13:23:56 Temperature Measurement, Probe Number = 1, T = 29.257000 C 13:23:56 Temperature Measurement, Probe Number = 2, T = 28.797000 C 13:23:57 Temperature Measurement, Probe Number = 3, T = 28.888000 C 13:23:57 Temperature Measurement, Probe Number = 4, T = 29.001000 C 13:23:58 Temperature Measurement, Probe Number = 5, T = 26.238000 C 13:23:58 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.49 P3 = RTD Outside, Value = 29.26 P4 = Supply Water, Value = 28.80 P5 = Return Water, Value = 28.89 P6 = @ RTD Outside, Value = 29.00 P7 = Ambient, Value = 26.24 13:23:58 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000040, Bz = -0.000020 T 13:24:06 Move Stage, position = 0.284000 13:24:07 Stage Position Measurement, pos = 0.2842800 m 13:24:08 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:24:09 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:24:10 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:24:10 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:24:10 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 13:24:11 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:24:12 Main Volt Voltage Measurement, V = 0.000002 V 13:24:12 Temperature Measurement, Probe Number = 0, T = 29.487000 C 13:24:16 Temperature Measurement, Probe Number = 1, T = 29.256000 C 13:24:16 Temperature Measurement, Probe Number = 2, T = 28.738000 C 13:24:17 Temperature Measurement, Probe Number = 3, T = 28.876000 C 13:24:18 Temperature Measurement, Probe Number = 4, T = 28.999000 C 13:24:18 Temperature Measurement, Probe Number = 5, T = 26.246000 C 13:24:18 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.49 P3 = RTD Outside, Value = 29.26 P4 = Supply Water, Value = 28.74 P5 = Return Water, Value = 28.88 P6 = @ RTD Outside, Value = 29.00 P7 = Ambient, Value = 26.25 13:24:18 3 Axis Hall Probe Field Measurement, Bx = 0.000010, By = 0.000050, Bz = -0.000030 T 13:24:26 Move Stage, position = 0.286000 13:24:28 Stage Position Measurement, pos = 0.2863100 m 13:24:28 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:24:29 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:24:30 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:24:30 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:24:30 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 13:24:31 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:24:32 Main Volt Voltage Measurement, V = 0.000002 V 13:24:32 Temperature Measurement, Probe Number = 0, T = 29.483000 C 13:24:36 Temperature Measurement, Probe Number = 1, T = 29.261000 C 13:24:37 Temperature Measurement, Probe Number = 2, T = 28.739000 C 13:24:37 Temperature Measurement, Probe Number = 3, T = 28.874000 C 13:24:38 Temperature Measurement, Probe Number = 4, T = 29.002000 C 13:24:38 Temperature Measurement, Probe Number = 5, T = 26.436000 C 13:24:38 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.48 P3 = RTD Outside, Value = 29.26 P4 = Supply Water, Value = 28.74 P5 = Return Water, Value = 28.87 P6 = @ RTD Outside, Value = 29.00 P7 = Ambient, Value = 26.44 13:24:38 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000040, Bz = -0.000030 T 13:24:46 Move Stage, position = 0.288000 13:24:48 Stage Position Measurement, pos = 0.2883050 m 13:24:48 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:24:49 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:24:50 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:24:51 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:24:51 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 13:24:51 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:24:52 Main Volt Voltage Measurement, V = 0.000001 V 13:24:52 Temperature Measurement, Probe Number = 0, T = 29.482000 C 13:24:56 Temperature Measurement, Probe Number = 1, T = 29.242000 C 13:24:57 Temperature Measurement, Probe Number = 2, T = 28.757000 C 13:24:57 Temperature Measurement, Probe Number = 3, T = 28.883000 C 13:24:58 Temperature Measurement, Probe Number = 4, T = 28.988000 C 13:24:58 Temperature Measurement, Probe Number = 5, T = 26.552000 C 13:24:58 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.48 P3 = RTD Outside, Value = 29.24 P4 = Supply Water, Value = 28.76 P5 = Return Water, Value = 28.88 P6 = @ RTD Outside, Value = 28.99 P7 = Ambient, Value = 26.55 13:24:58 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000050, Bz = -0.000020 T 13:25:06 Move Stage, position = 0.290000 13:25:08 Stage Position Measurement, pos = 0.2903150 m 13:25:08 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:25:09 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:25:10 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:25:11 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:25:11 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 13:25:11 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:25:12 Main Volt Voltage Measurement, V = 0.000001 V 13:25:13 Temperature Measurement, Probe Number = 0, T = 29.483000 C 13:25:16 Temperature Measurement, Probe Number = 1, T = 29.237000 C 13:25:17 Temperature Measurement, Probe Number = 2, T = 28.762000 C 13:25:17 Temperature Measurement, Probe Number = 3, T = 28.903000 C 13:25:18 Temperature Measurement, Probe Number = 4, T = 28.977000 C 13:25:18 Temperature Measurement, Probe Number = 5, T = 26.521000 C 13:25:18 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.48 P3 = RTD Outside, Value = 29.24 P4 = Supply Water, Value = 28.76 P5 = Return Water, Value = 28.90 P6 = @ RTD Outside, Value = 28.98 P7 = Ambient, Value = 26.52 13:25:19 3 Axis Hall Probe Field Measurement, Bx = -0.000010, By = 0.000040, Bz = -0.000020 T 13:25:26 Move Stage, position = 0.292000 13:25:28 Stage Position Measurement, pos = 0.2923150 m 13:25:29 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:25:29 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:25:30 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:25:31 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:25:31 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 13:25:31 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:25:32 Main Volt Voltage Measurement, V = 0.000001 V 13:25:33 Temperature Measurement, Probe Number = 0, T = 29.484000 C 13:25:36 Temperature Measurement, Probe Number = 1, T = 29.242000 C 13:25:37 Temperature Measurement, Probe Number = 2, T = 28.768000 C 13:25:37 Temperature Measurement, Probe Number = 3, T = 28.906000 C 13:25:38 Temperature Measurement, Probe Number = 4, T = 28.969000 C 13:25:39 Temperature Measurement, Probe Number = 5, T = 26.424000 C 13:25:39 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.48 P3 = RTD Outside, Value = 29.24 P4 = Supply Water, Value = 28.77 P5 = Return Water, Value = 28.91 P6 = @ RTD Outside, Value = 28.97 P7 = Ambient, Value = 26.42 13:25:39 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000030, Bz = -0.000030 T 13:25:46 Move Stage, position = 0.294000 13:25:48 Stage Position Measurement, pos = 0.2943350 m 13:25:49 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:25:49 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:25:50 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:25:51 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:25:51 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 13:25:52 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:25:52 Main Volt Voltage Measurement, V = 0.000001 V 13:25:53 Temperature Measurement, Probe Number = 0, T = 29.486000 C 13:25:56 Temperature Measurement, Probe Number = 1, T = 29.249000 C 13:25:57 Temperature Measurement, Probe Number = 2, T = 28.775000 C 13:25:58 Temperature Measurement, Probe Number = 3, T = 28.909000 C 13:25:58 Temperature Measurement, Probe Number = 4, T = 28.976000 C 13:25:59 Temperature Measurement, Probe Number = 5, T = 26.402000 C 13:25:59 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.49 P3 = RTD Outside, Value = 29.25 P4 = Supply Water, Value = 28.77 P5 = Return Water, Value = 28.91 P6 = @ RTD Outside, Value = 28.98 P7 = Ambient, Value = 26.40 13:25:59 3 Axis Hall Probe Field Measurement, Bx = -0.000010, By = 0.000040, Bz = -0.000020 T 13:26:06 Move Stage, position = 0.296000 13:26:08 Stage Position Measurement, pos = 0.2963700 m 13:26:09 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:26:10 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:26:10 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:26:11 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:26:11 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 13:26:12 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:26:12 Main Volt Voltage Measurement, V = 0.000001 V 13:26:13 Temperature Measurement, Probe Number = 0, T = 29.484000 C 13:26:17 Temperature Measurement, Probe Number = 1, T = 29.254000 C 13:26:17 Temperature Measurement, Probe Number = 2, T = 28.744000 C 13:26:18 Temperature Measurement, Probe Number = 3, T = 28.907000 C 13:26:18 Temperature Measurement, Probe Number = 4, T = 28.985000 C 13:26:19 Temperature Measurement, Probe Number = 5, T = 26.365000 C 13:26:19 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.48 P3 = RTD Outside, Value = 29.25 P4 = Supply Water, Value = 28.74 P5 = Return Water, Value = 28.91 P6 = @ RTD Outside, Value = 28.98 P7 = Ambient, Value = 26.36 13:26:19 3 Axis Hall Probe Field Measurement, Bx = -0.000010, By = 0.000040, Bz = -0.000030 T 13:26:27 Move Stage, position = 0.298000 13:26:28 Stage Position Measurement, pos = 0.2983300 m 13:26:29 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:26:30 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:26:30 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:26:31 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:26:31 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 13:26:32 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:26:32 Main Volt Voltage Measurement, V = 0.000001 V 13:26:33 Temperature Measurement, Probe Number = 0, T = 29.480000 C 13:26:37 Temperature Measurement, Probe Number = 1, T = 29.251000 C 13:26:37 Temperature Measurement, Probe Number = 2, T = 28.650000 C 13:26:38 Temperature Measurement, Probe Number = 3, T = 28.875000 C 13:26:38 Temperature Measurement, Probe Number = 4, T = 28.989000 C 13:26:39 Temperature Measurement, Probe Number = 5, T = 26.184000 C 13:26:39 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.48 P3 = RTD Outside, Value = 29.25 P4 = Supply Water, Value = 28.65 P5 = Return Water, Value = 28.88 P6 = @ RTD Outside, Value = 28.99 P7 = Ambient, Value = 26.18 13:26:39 3 Axis Hall Probe Field Measurement, Bx = -0.000010, By = 0.000040, Bz = -0.000020 T 13:26:47 Move Stage, position = 0.300000 13:26:48 Stage Position Measurement, pos = 0.3003300 m 13:26:49 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:26:50 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:26:50 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:26:51 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:26:51 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 13:26:52 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:26:52 Main Volt Voltage Measurement, V = 0.000002 V 13:26:53 Temperature Measurement, Probe Number = 0, T = 29.478000 C 13:26:57 Temperature Measurement, Probe Number = 1, T = 29.251000 C 13:26:57 Temperature Measurement, Probe Number = 2, T = 28.603000 C 13:26:58 Temperature Measurement, Probe Number = 3, T = 28.859000 C 13:26:58 Temperature Measurement, Probe Number = 4, T = 28.993000 C 13:26:59 Temperature Measurement, Probe Number = 5, T = 26.143000 C 13:26:59 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.48 P3 = RTD Outside, Value = 29.25 P4 = Supply Water, Value = 28.60 P5 = Return Water, Value = 28.86 P6 = @ RTD Outside, Value = 28.99 P7 = Ambient, Value = 26.14 13:26:59 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000040, Bz = -0.000030 T 13:27:07 Move Stage, position = 0.302000 13:27:08 Stage Position Measurement, pos = 0.3023050 m 13:27:09 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:27:10 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:27:11 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:27:11 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:27:11 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 13:27:12 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:27:13 Main Volt Voltage Measurement, V = 0.000002 V 13:27:13 Temperature Measurement, Probe Number = 0, T = 29.475000 C 13:27:17 Temperature Measurement, Probe Number = 1, T = 29.250000 C 13:27:17 Temperature Measurement, Probe Number = 2, T = 28.596000 C 13:27:18 Temperature Measurement, Probe Number = 3, T = 28.851000 C 13:27:19 Temperature Measurement, Probe Number = 4, T = 28.996000 C 13:27:19 Temperature Measurement, Probe Number = 5, T = 26.156000 C 13:27:19 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.48 P3 = RTD Outside, Value = 29.25 P4 = Supply Water, Value = 28.60 P5 = Return Water, Value = 28.85 P6 = @ RTD Outside, Value = 29.00 P7 = Ambient, Value = 26.16 13:27:19 3 Axis Hall Probe Field Measurement, Bx = -0.000010, By = 0.000040, Bz = -0.000020 T 13:27:27 Move Stage, position = 0.304000 13:27:29 Stage Position Measurement, pos = 0.3043150 m 13:27:29 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:27:30 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:27:31 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:27:31 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:27:31 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 13:27:32 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:27:33 Main Volt Voltage Measurement, V = 0.000002 V 13:27:33 Temperature Measurement, Probe Number = 0, T = 29.475000 C 13:27:37 Temperature Measurement, Probe Number = 1, T = 29.253000 C 13:27:38 Temperature Measurement, Probe Number = 2, T = 28.595000 C 13:27:38 Temperature Measurement, Probe Number = 3, T = 28.851000 C 13:27:39 Temperature Measurement, Probe Number = 4, T = 28.999000 C 13:27:39 Temperature Measurement, Probe Number = 5, T = 26.210000 C 13:27:39 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.48 P3 = RTD Outside, Value = 29.25 P4 = Supply Water, Value = 28.59 P5 = Return Water, Value = 28.85 P6 = @ RTD Outside, Value = 29.00 P7 = Ambient, Value = 26.21 13:27:39 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000040, Bz = -0.000020 T 13:27:47 Move Stage, position = 0.306000 13:27:49 Stage Position Measurement, pos = 0.3063150 m 13:27:49 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:27:50 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:27:51 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:27:52 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:27:52 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 13:27:52 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:27:53 Main Volt Voltage Measurement, V = 0.000002 V 13:27:54 Temperature Measurement, Probe Number = 0, T = 29.475000 C 13:27:57 Temperature Measurement, Probe Number = 1, T = 29.253000 C 13:27:58 Temperature Measurement, Probe Number = 2, T = 28.573000 C 13:27:58 Temperature Measurement, Probe Number = 3, T = 28.850000 C 13:27:59 Temperature Measurement, Probe Number = 4, T = 29.001000 C 13:27:59 Temperature Measurement, Probe Number = 5, T = 26.184000 C 13:27:59 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.48 P3 = RTD Outside, Value = 29.25 P4 = Supply Water, Value = 28.57 P5 = Return Water, Value = 28.85 P6 = @ RTD Outside, Value = 29.00 P7 = Ambient, Value = 26.18 13:28:00 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000040, Bz = -0.000020 T 13:28:07 Move Stage, position = 0.308000 13:28:09 Stage Position Measurement, pos = 0.3083550 m 13:28:10 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:28:10 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:28:11 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:28:12 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:28:12 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 13:28:12 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:28:13 Main Volt Voltage Measurement, V = 0.000002 V 13:28:14 Temperature Measurement, Probe Number = 0, T = 29.474000 C 13:28:17 Temperature Measurement, Probe Number = 1, T = 29.254000 C 13:28:18 Temperature Measurement, Probe Number = 2, T = 28.567000 C 13:28:18 Temperature Measurement, Probe Number = 3, T = 28.845000 C 13:28:19 Temperature Measurement, Probe Number = 4, T = 29.004000 C 13:28:20 Temperature Measurement, Probe Number = 5, T = 26.165000 C 13:28:20 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.47 P3 = RTD Outside, Value = 29.25 P4 = Supply Water, Value = 28.57 P5 = Return Water, Value = 28.84 P6 = @ RTD Outside, Value = 29.00 P7 = Ambient, Value = 26.16 13:28:20 3 Axis Hall Probe Field Measurement, Bx = -0.000010, By = 0.000040, Bz = -0.000020 T 13:28:27 Move Stage, position = 0.310000 13:28:29 Stage Position Measurement, pos = 0.3103600 m 13:28:30 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:28:30 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:28:31 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:28:32 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:28:32 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 13:28:33 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:28:33 Main Volt Voltage Measurement, V = 0.000002 V 13:28:34 Temperature Measurement, Probe Number = 0, T = 29.473000 C 13:28:37 Temperature Measurement, Probe Number = 1, T = 29.252000 C 13:28:38 Temperature Measurement, Probe Number = 2, T = 28.587000 C 13:28:39 Temperature Measurement, Probe Number = 3, T = 28.840000 C 13:28:39 Temperature Measurement, Probe Number = 4, T = 29.007000 C 13:28:40 Temperature Measurement, Probe Number = 5, T = 26.132000 C 13:28:40 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.47 P3 = RTD Outside, Value = 29.25 P4 = Supply Water, Value = 28.59 P5 = Return Water, Value = 28.84 P6 = @ RTD Outside, Value = 29.01 P7 = Ambient, Value = 26.13 13:28:40 3 Axis Hall Probe Field Measurement, Bx = -0.000010, By = 0.000040, Bz = -0.000020 T 13:28:48 Move Stage, position = 0.312000 13:28:49 Stage Position Measurement, pos = 0.3123600 m 13:28:50 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:28:51 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:28:51 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:28:52 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:28:52 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 13:28:53 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:28:53 Main Volt Voltage Measurement, V = 0.000002 V 13:28:54 Temperature Measurement, Probe Number = 0, T = 29.474000 C 13:28:58 Temperature Measurement, Probe Number = 1, T = 29.254000 C 13:28:58 Temperature Measurement, Probe Number = 2, T = 28.597000 C 13:28:59 Temperature Measurement, Probe Number = 3, T = 28.846000 C 13:28:59 Temperature Measurement, Probe Number = 4, T = 29.009000 C 13:29:00 Temperature Measurement, Probe Number = 5, T = 26.106000 C 13:29:00 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.47 P3 = RTD Outside, Value = 29.25 P4 = Supply Water, Value = 28.60 P5 = Return Water, Value = 28.85 P6 = @ RTD Outside, Value = 29.01 P7 = Ambient, Value = 26.11 13:29:00 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000040, Bz = -0.000020 T 13:29:08 Move Stage, position = 0.314000 13:29:09 Stage Position Measurement, pos = 0.3143800 m 13:29:10 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:29:11 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:29:12 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:29:12 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:29:12 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 13:29:13 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:29:14 Main Volt Voltage Measurement, V = 0.000002 V 13:29:14 Temperature Measurement, Probe Number = 0, T = 29.472000 C 13:29:18 Temperature Measurement, Probe Number = 1, T = 29.249000 C 13:29:18 Temperature Measurement, Probe Number = 2, T = 28.601000 C 13:29:19 Temperature Measurement, Probe Number = 3, T = 28.857000 C 13:29:19 Temperature Measurement, Probe Number = 4, T = 29.009000 C 13:29:20 Temperature Measurement, Probe Number = 5, T = 26.110000 C 13:29:20 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.47 P3 = RTD Outside, Value = 29.25 P4 = Supply Water, Value = 28.60 P5 = Return Water, Value = 28.86 P6 = @ RTD Outside, Value = 29.01 P7 = Ambient, Value = 26.11 13:29:20 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000040, Bz = -0.000020 T 13:29:28 Move Stage, position = 0.316000 13:29:29 Stage Position Measurement, pos = 0.3163650 m 13:29:30 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:29:31 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:29:32 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:29:32 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:29:32 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 13:29:33 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:29:34 Main Volt Voltage Measurement, V = 0.000002 V 13:29:34 Temperature Measurement, Probe Number = 0, T = 29.469000 C 13:29:38 Temperature Measurement, Probe Number = 1, T = 29.245000 C 13:29:38 Temperature Measurement, Probe Number = 2, T = 28.624000 C 13:29:39 Temperature Measurement, Probe Number = 3, T = 28.854000 C 13:29:40 Temperature Measurement, Probe Number = 4, T = 29.007000 C 13:29:40 Temperature Measurement, Probe Number = 5, T = 26.156000 C 13:29:40 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.47 P3 = RTD Outside, Value = 29.25 P4 = Supply Water, Value = 28.62 P5 = Return Water, Value = 28.85 P6 = @ RTD Outside, Value = 29.01 P7 = Ambient, Value = 26.16 13:29:40 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000040, Bz = -0.000020 T 13:29:48 Move Stage, position = 0.318000 13:29:50 Stage Position Measurement, pos = 0.3183700 m 13:29:50 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:29:51 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:29:52 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:29:52 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:29:52 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 13:29:53 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:29:54 Main Volt Voltage Measurement, V = 0.000002 V 13:29:54 Temperature Measurement, Probe Number = 0, T = 29.469000 C 13:29:58 Temperature Measurement, Probe Number = 1, T = 29.245000 C 13:29:59 Temperature Measurement, Probe Number = 2, T = 28.649000 C 13:29:59 Temperature Measurement, Probe Number = 3, T = 28.839000 C 13:30:00 Temperature Measurement, Probe Number = 4, T = 29.005000 C 13:30:00 Temperature Measurement, Probe Number = 5, T = 26.177000 C 13:30:00 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.47 P3 = RTD Outside, Value = 29.25 P4 = Supply Water, Value = 28.65 P5 = Return Water, Value = 28.84 P6 = @ RTD Outside, Value = 29.00 P7 = Ambient, Value = 26.18 13:30:00 3 Axis Hall Probe Field Measurement, Bx = 0.000010, By = 0.000040, Bz = -0.000020 T 13:30:08 Move Stage, position = 0.320000 13:30:10 Stage Position Measurement, pos = 0.3203500 m 13:30:10 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:30:11 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:30:12 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:30:13 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:30:13 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 13:30:13 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:30:14 Main Volt Voltage Measurement, V = 0.000002 V 13:30:15 Temperature Measurement, Probe Number = 0, T = 29.466000 C 13:30:18 Temperature Measurement, Probe Number = 1, T = 29.240000 C 13:30:19 Temperature Measurement, Probe Number = 2, T = 28.667000 C 13:30:19 Temperature Measurement, Probe Number = 3, T = 28.836000 C 13:30:20 Temperature Measurement, Probe Number = 4, T = 29.001000 C 13:30:20 Temperature Measurement, Probe Number = 5, T = 26.360000 C 13:30:20 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.47 P3 = RTD Outside, Value = 29.24 P4 = Supply Water, Value = 28.67 P5 = Return Water, Value = 28.84 P6 = @ RTD Outside, Value = 29.00 P7 = Ambient, Value = 26.36 13:30:21 3 Axis Hall Probe Field Measurement, Bx = -0.000010, By = 0.000040, Bz = -0.000020 T 13:30:28 Move Stage, position = 0.322000 13:30:30 Stage Position Measurement, pos = 0.3223500 m 13:30:31 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:30:31 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:30:32 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:30:33 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:30:33 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 13:30:33 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:30:34 Main Volt Voltage Measurement, V = 0.000002 V 13:30:35 Temperature Measurement, Probe Number = 0, T = 29.463000 C 13:30:38 Temperature Measurement, Probe Number = 1, T = 29.210000 C 13:30:39 Temperature Measurement, Probe Number = 2, T = 28.678000 C 13:30:39 Temperature Measurement, Probe Number = 3, T = 28.843000 C 13:30:40 Temperature Measurement, Probe Number = 4, T = 28.968000 C 13:30:41 Temperature Measurement, Probe Number = 5, T = 26.435000 C 13:30:41 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.46 P3 = RTD Outside, Value = 29.21 P4 = Supply Water, Value = 28.68 P5 = Return Water, Value = 28.84 P6 = @ RTD Outside, Value = 28.97 P7 = Ambient, Value = 26.43 13:30:41 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000040, Bz = -0.000030 T 13:30:48 Move Stage, position = 0.324000 13:30:50 Stage Position Measurement, pos = 0.3243400 m 13:30:51 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:30:51 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:30:52 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:30:53 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:30:53 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 13:30:54 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:30:54 Main Volt Voltage Measurement, V = 0.000002 V 13:30:55 Temperature Measurement, Probe Number = 0, T = 29.460000 C 13:30:58 Temperature Measurement, Probe Number = 1, T = 29.189000 C 13:30:59 Temperature Measurement, Probe Number = 2, T = 28.659000 C 13:31:00 Temperature Measurement, Probe Number = 3, T = 28.847000 C 13:31:00 Temperature Measurement, Probe Number = 4, T = 28.926000 C 13:31:01 Temperature Measurement, Probe Number = 5, T = 26.362000 C 13:31:01 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.46 P3 = RTD Outside, Value = 29.19 P4 = Supply Water, Value = 28.66 P5 = Return Water, Value = 28.85 P6 = @ RTD Outside, Value = 28.93 P7 = Ambient, Value = 26.36 13:31:01 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000040, Bz = -0.000030 T 13:31:09 Move Stage, position = 0.326000 13:31:10 Stage Position Measurement, pos = 0.3263650 m 13:31:11 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:31:12 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:31:12 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:31:13 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:31:13 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 13:31:14 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:31:14 Main Volt Voltage Measurement, V = 0.000002 V 13:31:15 Temperature Measurement, Probe Number = 0, T = 29.460000 C 13:31:19 Temperature Measurement, Probe Number = 1, T = 29.203000 C 13:31:19 Temperature Measurement, Probe Number = 2, T = 28.635000 C 13:31:20 Temperature Measurement, Probe Number = 3, T = 28.847000 C 13:31:20 Temperature Measurement, Probe Number = 4, T = 28.922000 C 13:31:21 Temperature Measurement, Probe Number = 5, T = 26.339000 C 13:31:21 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.46 P3 = RTD Outside, Value = 29.20 P4 = Supply Water, Value = 28.64 P5 = Return Water, Value = 28.85 P6 = @ RTD Outside, Value = 28.92 P7 = Ambient, Value = 26.34 13:31:21 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000040, Bz = -0.000020 T 13:31:29 Move Stage, position = 0.328000 13:31:30 Stage Position Measurement, pos = 0.3283650 m 13:31:31 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:31:32 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:31:32 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:31:33 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:31:33 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 13:31:34 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:31:34 Main Volt Voltage Measurement, V = 0.000002 V 13:31:35 Temperature Measurement, Probe Number = 0, T = 29.457000 C 13:31:39 Temperature Measurement, Probe Number = 1, T = 29.216000 C 13:31:39 Temperature Measurement, Probe Number = 2, T = 28.647000 C 13:31:40 Temperature Measurement, Probe Number = 3, T = 28.835000 C 13:31:40 Temperature Measurement, Probe Number = 4, T = 28.930000 C 13:31:41 Temperature Measurement, Probe Number = 5, T = 26.274000 C 13:31:41 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.46 P3 = RTD Outside, Value = 29.22 P4 = Supply Water, Value = 28.65 P5 = Return Water, Value = 28.84 P6 = @ RTD Outside, Value = 28.93 P7 = Ambient, Value = 26.27 13:31:41 3 Axis Hall Probe Field Measurement, Bx = -0.000010, By = 0.000050, Bz = -0.000020 T 13:31:49 Move Stage, position = 0.330000 13:31:50 Stage Position Measurement, pos = 0.3303850 m 13:31:51 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:31:52 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:31:53 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:31:53 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:31:53 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 13:31:54 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:31:55 Main Volt Voltage Measurement, V = 0.000002 V 13:31:55 Temperature Measurement, Probe Number = 0, T = 29.460000 C 13:31:59 Temperature Measurement, Probe Number = 1, T = 29.223000 C 13:31:59 Temperature Measurement, Probe Number = 2, T = 28.665000 C 13:32:00 Temperature Measurement, Probe Number = 3, T = 28.808000 C 13:32:01 Temperature Measurement, Probe Number = 4, T = 28.940000 C 13:32:01 Temperature Measurement, Probe Number = 5, T = 26.258000 C 13:32:01 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.46 P3 = RTD Outside, Value = 29.22 P4 = Supply Water, Value = 28.66 P5 = Return Water, Value = 28.81 P6 = @ RTD Outside, Value = 28.94 P7 = Ambient, Value = 26.26 13:32:01 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000040, Bz = -0.000020 T 13:32:09 Move Stage, position = 0.332000 13:32:10 Stage Position Measurement, pos = 0.3323900 m 13:32:11 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:32:12 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:32:13 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:32:13 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:32:13 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 13:32:14 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:32:15 Main Volt Voltage Measurement, V = 0.000002 V 13:32:15 Temperature Measurement, Probe Number = 0, T = 29.459000 C 13:32:19 Temperature Measurement, Probe Number = 1, T = 29.228000 C 13:32:19 Temperature Measurement, Probe Number = 2, T = 28.681000 C 13:32:20 Temperature Measurement, Probe Number = 3, T = 28.806000 C 13:32:21 Temperature Measurement, Probe Number = 4, T = 28.944000 C 13:32:21 Temperature Measurement, Probe Number = 5, T = 26.252000 C 13:32:21 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.46 P3 = RTD Outside, Value = 29.23 P4 = Supply Water, Value = 28.68 P5 = Return Water, Value = 28.81 P6 = @ RTD Outside, Value = 28.94 P7 = Ambient, Value = 26.25 13:32:21 3 Axis Hall Probe Field Measurement, Bx = -0.000010, By = 0.000040, Bz = -0.000030 T 13:32:29 Move Stage, position = 0.334000 13:32:31 Stage Position Measurement, pos = 0.3343650 m 13:32:31 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:32:32 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:32:33 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:32:33 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:32:33 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 13:32:34 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:32:35 Main Volt Voltage Measurement, V = 0.000002 V 13:32:35 Temperature Measurement, Probe Number = 0, T = 29.460000 C 13:32:39 Temperature Measurement, Probe Number = 1, T = 29.214000 C 13:32:40 Temperature Measurement, Probe Number = 2, T = 28.707000 C 13:32:40 Temperature Measurement, Probe Number = 3, T = 28.786000 C 13:32:41 Temperature Measurement, Probe Number = 4, T = 28.939000 C 13:32:41 Temperature Measurement, Probe Number = 5, T = 26.299000 C 13:32:41 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.46 P3 = RTD Outside, Value = 29.21 P4 = Supply Water, Value = 28.71 P5 = Return Water, Value = 28.79 P6 = @ RTD Outside, Value = 28.94 P7 = Ambient, Value = 26.30 13:32:41 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000040, Bz = -0.000020 T 13:32:49 Move Stage, position = 0.336000 13:32:51 Stage Position Measurement, pos = 0.3363850 m 13:32:51 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:32:52 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:32:53 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:32:53 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:32:53 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 13:32:54 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:32:55 Main Volt Voltage Measurement, V = 0.000002 V 13:32:55 Temperature Measurement, Probe Number = 0, T = 29.462000 C 13:32:59 Temperature Measurement, Probe Number = 1, T = 29.228000 C 13:33:00 Temperature Measurement, Probe Number = 2, T = 28.724000 C 13:33:00 Temperature Measurement, Probe Number = 3, T = 28.778000 C 13:33:01 Temperature Measurement, Probe Number = 4, T = 28.946000 C 13:33:01 Temperature Measurement, Probe Number = 5, T = 26.369000 C 13:33:01 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.46 P3 = RTD Outside, Value = 29.23 P4 = Supply Water, Value = 28.72 P5 = Return Water, Value = 28.78 P6 = @ RTD Outside, Value = 28.95 P7 = Ambient, Value = 26.37 13:33:01 3 Axis Hall Probe Field Measurement, Bx = -0.000010, By = 0.000050, Bz = -0.000030 T 13:33:09 Move Stage, position = 0.338000 13:33:11 Stage Position Measurement, pos = 0.3383600 m 13:33:11 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:33:12 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:33:13 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:33:14 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:33:14 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 13:33:14 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:33:15 Main Volt Voltage Measurement, V = 0.000002 V 13:33:15 Temperature Measurement, Probe Number = 0, T = 29.465000 C 13:33:19 Temperature Measurement, Probe Number = 1, T = 29.231000 C 13:33:20 Temperature Measurement, Probe Number = 2, T = 28.744000 C 13:33:20 Temperature Measurement, Probe Number = 3, T = 28.762000 C 13:33:21 Temperature Measurement, Probe Number = 4, T = 28.954000 C 13:33:21 Temperature Measurement, Probe Number = 5, T = 26.471000 C 13:33:21 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.46 P3 = RTD Outside, Value = 29.23 P4 = Supply Water, Value = 28.74 P5 = Return Water, Value = 28.76 P6 = @ RTD Outside, Value = 28.95 P7 = Ambient, Value = 26.47 13:33:22 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000040, Bz = -0.000030 T 13:33:29 Move Stage, position = 0.340000 13:33:31 Stage Position Measurement, pos = 0.3403650 m 13:33:32 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:33:32 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:33:33 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:33:34 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:33:34 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 13:33:34 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:33:35 Main Volt Voltage Measurement, V = 0.000002 V 13:33:36 Temperature Measurement, Probe Number = 0, T = 29.469000 C 13:33:39 Temperature Measurement, Probe Number = 1, T = 29.233000 C 13:33:40 Temperature Measurement, Probe Number = 2, T = 28.766000 C 13:33:40 Temperature Measurement, Probe Number = 3, T = 28.786000 C 13:33:41 Temperature Measurement, Probe Number = 4, T = 28.957000 C 13:33:42 Temperature Measurement, Probe Number = 5, T = 26.416000 C 13:33:42 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.47 P3 = RTD Outside, Value = 29.23 P4 = Supply Water, Value = 28.77 P5 = Return Water, Value = 28.79 P6 = @ RTD Outside, Value = 28.96 P7 = Ambient, Value = 26.42 13:33:42 3 Axis Hall Probe Field Measurement, Bx = -0.000010, By = 0.000040, Bz = -0.000010 T 13:33:49 Move Stage, position = 0.342000 13:33:51 Stage Position Measurement, pos = 0.3423500 m 13:33:52 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:33:52 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:33:53 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:33:54 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:33:54 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 13:33:55 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:33:55 Main Volt Voltage Measurement, V = 0.000002 V 13:33:56 Temperature Measurement, Probe Number = 0, T = 29.470000 C 13:33:59 Temperature Measurement, Probe Number = 1, T = 29.239000 C 13:34:00 Temperature Measurement, Probe Number = 2, T = 28.780000 C 13:34:01 Temperature Measurement, Probe Number = 3, T = 28.822000 C 13:34:01 Temperature Measurement, Probe Number = 4, T = 28.969000 C 13:34:02 Temperature Measurement, Probe Number = 5, T = 26.374000 C 13:34:02 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.47 P3 = RTD Outside, Value = 29.24 P4 = Supply Water, Value = 28.78 P5 = Return Water, Value = 28.82 P6 = @ RTD Outside, Value = 28.97 P7 = Ambient, Value = 26.37 13:34:02 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000030, Bz = -0.000030 T 13:34:09 Move Stage, position = 0.344000 13:34:11 Stage Position Measurement, pos = 0.3443650 m 13:34:12 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:34:12 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:34:13 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:34:14 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:34:14 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 13:34:15 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:34:15 Main Volt Voltage Measurement, V = 0.000002 V 13:34:16 Temperature Measurement, Probe Number = 0, T = 29.467000 C 13:34:19 Temperature Measurement, Probe Number = 1, T = 29.241000 C 13:34:20 Temperature Measurement, Probe Number = 2, T = 28.779000 C 13:34:21 Temperature Measurement, Probe Number = 3, T = 28.839000 C 13:34:21 Temperature Measurement, Probe Number = 4, T = 28.977000 C 13:34:22 Temperature Measurement, Probe Number = 5, T = 26.386000 C 13:34:22 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.47 P3 = RTD Outside, Value = 29.24 P4 = Supply Water, Value = 28.78 P5 = Return Water, Value = 28.84 P6 = @ RTD Outside, Value = 28.98 P7 = Ambient, Value = 26.39 13:34:22 3 Axis Hall Probe Field Measurement, Bx = -0.000010, By = 0.000040, Bz = -0.000030 T 13:34:30 Move Stage, position = 0.346000 13:34:31 Stage Position Measurement, pos = 0.3463950 m 13:34:32 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:34:33 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:34:33 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:34:34 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:34:34 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 13:34:35 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:34:35 Main Volt Voltage Measurement, V = 0.000002 V 13:34:36 Temperature Measurement, Probe Number = 0, T = 29.468000 C 13:34:40 Temperature Measurement, Probe Number = 1, T = 29.244000 C 13:34:40 Temperature Measurement, Probe Number = 2, T = 28.790000 C 13:34:41 Temperature Measurement, Probe Number = 3, T = 28.824000 C 13:34:41 Temperature Measurement, Probe Number = 4, T = 28.982000 C 13:34:42 Temperature Measurement, Probe Number = 5, T = 26.389000 C 13:34:42 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.47 P3 = RTD Outside, Value = 29.24 P4 = Supply Water, Value = 28.79 P5 = Return Water, Value = 28.82 P6 = @ RTD Outside, Value = 28.98 P7 = Ambient, Value = 26.39 13:34:42 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000040, Bz = -0.000030 T 13:34:50 Move Stage, position = 0.348000 13:34:51 Stage Position Measurement, pos = 0.3484000 m 13:34:52 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:34:53 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:34:53 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:34:54 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:34:54 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 13:34:55 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:34:55 Main Volt Voltage Measurement, V = 0.000002 V 13:34:56 Temperature Measurement, Probe Number = 0, T = 29.472000 C 13:35:00 Temperature Measurement, Probe Number = 1, T = 29.247000 C 13:35:00 Temperature Measurement, Probe Number = 2, T = 28.800000 C 13:35:01 Temperature Measurement, Probe Number = 3, T = 28.825000 C 13:35:01 Temperature Measurement, Probe Number = 4, T = 28.989000 C 13:35:02 Temperature Measurement, Probe Number = 5, T = 26.361000 C 13:35:02 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.47 P3 = RTD Outside, Value = 29.25 P4 = Supply Water, Value = 28.80 P5 = Return Water, Value = 28.82 P6 = @ RTD Outside, Value = 28.99 P7 = Ambient, Value = 26.36 13:35:02 3 Axis Hall Probe Field Measurement, Bx = 0.000010, By = 0.000030, Bz = -0.000020 T 13:35:10 Move Stage, position = 0.350000 13:35:11 Stage Position Measurement, pos = 0.3504050 m 13:35:12 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:35:13 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:35:14 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:35:14 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:35:14 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 13:35:15 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:35:16 Main Volt Voltage Measurement, V = 0.000002 V 13:35:16 Temperature Measurement, Probe Number = 0, T = 29.469000 C 13:35:20 Temperature Measurement, Probe Number = 1, T = 29.235000 C 13:35:20 Temperature Measurement, Probe Number = 2, T = 28.771000 C 13:35:21 Temperature Measurement, Probe Number = 3, T = 28.842000 C 13:35:22 Temperature Measurement, Probe Number = 4, T = 28.984000 C 13:35:22 Temperature Measurement, Probe Number = 5, T = 26.535000 C 13:35:22 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.47 P3 = RTD Outside, Value = 29.23 P4 = Supply Water, Value = 28.77 P5 = Return Water, Value = 28.84 P6 = @ RTD Outside, Value = 28.98 P7 = Ambient, Value = 26.54 13:35:22 3 Axis Hall Probe Field Measurement, Bx = -0.000010, By = 0.000040, Bz = -0.000020 T 13:35:30 Move Stage, position = 0.352000 13:35:31 Stage Position Measurement, pos = 0.3523900 m 13:35:32 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:35:33 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:35:34 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:35:34 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:35:34 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 13:35:35 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:35:36 Main Volt Voltage Measurement, V = 0.000002 V 13:35:36 Temperature Measurement, Probe Number = 0, T = 29.462000 C 13:35:40 Temperature Measurement, Probe Number = 1, T = 29.195000 C 13:35:41 Temperature Measurement, Probe Number = 2, T = 28.743000 C 13:35:41 Temperature Measurement, Probe Number = 3, T = 28.850000 C 13:35:42 Temperature Measurement, Probe Number = 4, T = 28.936000 C 13:35:42 Temperature Measurement, Probe Number = 5, T = 26.566000 C 13:35:42 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.46 P3 = RTD Outside, Value = 29.20 P4 = Supply Water, Value = 28.74 P5 = Return Water, Value = 28.85 P6 = @ RTD Outside, Value = 28.94 P7 = Ambient, Value = 26.57 13:35:42 3 Axis Hall Probe Field Measurement, Bx = -0.000010, By = 0.000040, Bz = -0.000030 T 13:35:50 Move Stage, position = 0.354000 13:35:52 Stage Position Measurement, pos = 0.3544100 m 13:35:53 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:35:53 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:35:54 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:35:55 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:35:55 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 13:35:55 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:35:56 Main Volt Voltage Measurement, V = 0.000002 V 13:35:57 Temperature Measurement, Probe Number = 0, T = 29.457000 C 13:36:00 Temperature Measurement, Probe Number = 1, T = 29.165000 C 13:36:01 Temperature Measurement, Probe Number = 2, T = 28.724000 C 13:36:01 Temperature Measurement, Probe Number = 3, T = 28.851000 C 13:36:02 Temperature Measurement, Probe Number = 4, T = 28.863000 C 13:36:03 Temperature Measurement, Probe Number = 5, T = 26.521000 C 13:36:03 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.46 P3 = RTD Outside, Value = 29.16 P4 = Supply Water, Value = 28.72 P5 = Return Water, Value = 28.85 P6 = @ RTD Outside, Value = 28.86 P7 = Ambient, Value = 26.52 13:36:03 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000040, Bz = -0.000020 T 13:36:10 Move Stage, position = 0.356000 13:36:12 Stage Position Measurement, pos = 0.3563950 m 13:36:13 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:36:13 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:36:14 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:36:15 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:36:15 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 13:36:15 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:36:16 Main Volt Voltage Measurement, V = 0.000002 V 13:36:17 Temperature Measurement, Probe Number = 0, T = 29.456000 C 13:36:20 Temperature Measurement, Probe Number = 1, T = 29.163000 C 13:36:21 Temperature Measurement, Probe Number = 2, T = 28.709000 C 13:36:21 Temperature Measurement, Probe Number = 3, T = 28.863000 C 13:36:22 Temperature Measurement, Probe Number = 4, T = 28.832000 C 13:36:23 Temperature Measurement, Probe Number = 5, T = 26.476000 C 13:36:23 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.46 P3 = RTD Outside, Value = 29.16 P4 = Supply Water, Value = 28.71 P5 = Return Water, Value = 28.86 P6 = @ RTD Outside, Value = 28.83 P7 = Ambient, Value = 26.48 13:36:23 3 Axis Hall Probe Field Measurement, Bx = -0.000010, By = 0.000040, Bz = -0.000030 T 13:36:30 Move Stage, position = 0.358000 13:36:32 Stage Position Measurement, pos = 0.3584200 m 13:36:33 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:36:33 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:36:34 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:36:35 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:36:35 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 13:36:36 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:36:36 Main Volt Voltage Measurement, V = 0.000002 V 13:36:37 Temperature Measurement, Probe Number = 0, T = 29.456000 C 13:36:40 Temperature Measurement, Probe Number = 1, T = 29.177000 C 13:36:41 Temperature Measurement, Probe Number = 2, T = 28.719000 C 13:36:42 Temperature Measurement, Probe Number = 3, T = 28.867000 C 13:36:42 Temperature Measurement, Probe Number = 4, T = 28.836000 C 13:36:43 Temperature Measurement, Probe Number = 5, T = 26.483000 C 13:36:43 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.46 P3 = RTD Outside, Value = 29.18 P4 = Supply Water, Value = 28.72 P5 = Return Water, Value = 28.87 P6 = @ RTD Outside, Value = 28.84 P7 = Ambient, Value = 26.48 13:36:43 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000040, Bz = -0.000020 T 13:36:50 Move Stage, position = 0.360000 13:36:52 Stage Position Measurement, pos = 0.3604050 m 13:36:53 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:36:54 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:36:54 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:36:55 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:36:55 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 13:36:56 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:36:56 Main Volt Voltage Measurement, V = 0.000002 V 13:36:57 Temperature Measurement, Probe Number = 0, T = 29.456000 C 13:37:01 Temperature Measurement, Probe Number = 1, T = 29.189000 C 13:37:01 Temperature Measurement, Probe Number = 2, T = 28.738000 C 13:37:02 Temperature Measurement, Probe Number = 3, T = 28.872000 C 13:37:02 Temperature Measurement, Probe Number = 4, T = 28.857000 C 13:37:03 Temperature Measurement, Probe Number = 5, T = 26.498000 C 13:37:03 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.46 P3 = RTD Outside, Value = 29.19 P4 = Supply Water, Value = 28.74 P5 = Return Water, Value = 28.87 P6 = @ RTD Outside, Value = 28.86 P7 = Ambient, Value = 26.50 13:37:03 3 Axis Hall Probe Field Measurement, Bx = -0.000010, By = 0.000040, Bz = -0.000020 T 13:37:11 Move Stage, position = 0.362000 13:37:12 Stage Position Measurement, pos = 0.3624200 m 13:37:13 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:37:14 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:37:14 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:37:15 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:37:15 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 13:37:16 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:37:16 Main Volt Voltage Measurement, V = 0.000002 V 13:37:17 Temperature Measurement, Probe Number = 0, T = 29.458000 C 13:37:21 Temperature Measurement, Probe Number = 1, T = 29.209000 C 13:37:21 Temperature Measurement, Probe Number = 2, T = 28.739000 C 13:37:22 Temperature Measurement, Probe Number = 3, T = 28.880000 C 13:37:22 Temperature Measurement, Probe Number = 4, T = 28.887000 C 13:37:23 Temperature Measurement, Probe Number = 5, T = 26.474000 C 13:37:23 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.46 P3 = RTD Outside, Value = 29.21 P4 = Supply Water, Value = 28.74 P5 = Return Water, Value = 28.88 P6 = @ RTD Outside, Value = 28.89 P7 = Ambient, Value = 26.47 13:37:23 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000040, Bz = -0.000020 T 13:37:31 Move Stage, position = 0.364000 13:37:32 Stage Position Measurement, pos = 0.3644550 m 13:37:33 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:37:34 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:37:35 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:37:35 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:37:35 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 13:37:36 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:37:37 Main Volt Voltage Measurement, V = 0.000001 V 13:37:37 Temperature Measurement, Probe Number = 0, T = 29.458000 C 13:37:41 Temperature Measurement, Probe Number = 1, T = 29.216000 C 13:37:41 Temperature Measurement, Probe Number = 2, T = 28.718000 C 13:37:42 Temperature Measurement, Probe Number = 3, T = 28.875000 C 13:37:42 Temperature Measurement, Probe Number = 4, T = 28.911000 C 13:37:43 Temperature Measurement, Probe Number = 5, T = 26.410000 C 13:37:43 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.46 P3 = RTD Outside, Value = 29.22 P4 = Supply Water, Value = 28.72 P5 = Return Water, Value = 28.88 P6 = @ RTD Outside, Value = 28.91 P7 = Ambient, Value = 26.41 13:37:43 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000050, Bz = -0.000030 T 13:37:51 Move Stage, position = 0.366000 13:37:52 Stage Position Measurement, pos = 0.3664850 m 13:37:53 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:37:54 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:37:55 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:37:55 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:37:55 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 13:37:56 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:37:57 Main Volt Voltage Measurement, V = 0.000002 V 13:37:57 Temperature Measurement, Probe Number = 0, T = 29.454000 C 13:38:01 Temperature Measurement, Probe Number = 1, T = 29.219000 C 13:38:01 Temperature Measurement, Probe Number = 2, T = 28.677000 C 13:38:02 Temperature Measurement, Probe Number = 3, T = 28.861000 C 13:38:03 Temperature Measurement, Probe Number = 4, T = 28.927000 C 13:38:03 Temperature Measurement, Probe Number = 5, T = 26.399000 C 13:38:03 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.45 P3 = RTD Outside, Value = 29.22 P4 = Supply Water, Value = 28.68 P5 = Return Water, Value = 28.86 P6 = @ RTD Outside, Value = 28.93 P7 = Ambient, Value = 26.40 13:38:03 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000040, Bz = -0.000020 T 13:38:11 Move Stage, position = 0.368000 13:38:13 Stage Position Measurement, pos = 0.3684900 m 13:38:13 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:38:14 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:38:15 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:38:15 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:38:15 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 13:38:16 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:38:17 Main Volt Voltage Measurement, V = 0.000002 V 13:38:17 Temperature Measurement, Probe Number = 0, T = 29.456000 C 13:38:21 Temperature Measurement, Probe Number = 1, T = 29.222000 C 13:38:22 Temperature Measurement, Probe Number = 2, T = 28.640000 C 13:38:22 Temperature Measurement, Probe Number = 3, T = 28.848000 C 13:38:23 Temperature Measurement, Probe Number = 4, T = 28.944000 C 13:38:23 Temperature Measurement, Probe Number = 5, T = 26.421000 C 13:38:23 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.46 P3 = RTD Outside, Value = 29.22 P4 = Supply Water, Value = 28.64 P5 = Return Water, Value = 28.85 P6 = @ RTD Outside, Value = 28.94 P7 = Ambient, Value = 26.42 13:38:23 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000040, Bz = -0.000020 T 13:38:31 Move Stage, position = 0.370000 13:38:33 Stage Position Measurement, pos = 0.3704700 m 13:38:33 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:38:34 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:38:35 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:38:36 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:38:36 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 13:38:36 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:38:37 Main Volt Voltage Measurement, V = 0.000002 V 13:38:37 Temperature Measurement, Probe Number = 0, T = 29.456000 C 13:38:41 Temperature Measurement, Probe Number = 1, T = 29.226000 C 13:38:42 Temperature Measurement, Probe Number = 2, T = 28.629000 C 13:38:42 Temperature Measurement, Probe Number = 3, T = 28.849000 C 13:38:43 Temperature Measurement, Probe Number = 4, T = 28.959000 C 13:38:44 Temperature Measurement, Probe Number = 5, T = 26.390000 C 13:38:44 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.46 P3 = RTD Outside, Value = 29.23 P4 = Supply Water, Value = 28.63 P5 = Return Water, Value = 28.85 P6 = @ RTD Outside, Value = 28.96 P7 = Ambient, Value = 26.39 13:38:44 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000050, Bz = -0.000020 T 13:38:51 Move Stage, position = 0.372000 13:38:53 Stage Position Measurement, pos = 0.3724850 m 13:38:54 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:38:54 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:38:55 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:38:56 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:38:56 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 13:38:57 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:38:57 Main Volt Voltage Measurement, V = 0.000002 V 13:38:58 Temperature Measurement, Probe Number = 0, T = 29.454000 C 13:39:01 Temperature Measurement, Probe Number = 1, T = 29.231000 C 13:39:02 Temperature Measurement, Probe Number = 2, T = 28.635000 C 13:39:03 Temperature Measurement, Probe Number = 3, T = 28.844000 C 13:39:03 Temperature Measurement, Probe Number = 4, T = 28.971000 C 13:39:04 Temperature Measurement, Probe Number = 5, T = 26.269000 C 13:39:04 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.45 P3 = RTD Outside, Value = 29.23 P4 = Supply Water, Value = 28.64 P5 = Return Water, Value = 28.84 P6 = @ RTD Outside, Value = 28.97 P7 = Ambient, Value = 26.27 13:39:04 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000040, Bz = -0.000010 T 13:39:11 Move Stage, position = 0.374000 13:39:13 Stage Position Measurement, pos = 0.3744800 m 13:39:14 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:39:15 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:39:15 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:39:16 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:39:16 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 13:39:17 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:39:17 Main Volt Voltage Measurement, V = 0.000002 V 13:39:18 Temperature Measurement, Probe Number = 0, T = 29.452000 C 13:39:21 Temperature Measurement, Probe Number = 1, T = 29.229000 C 13:39:22 Temperature Measurement, Probe Number = 2, T = 28.613000 C 13:39:23 Temperature Measurement, Probe Number = 3, T = 28.843000 C 13:39:23 Temperature Measurement, Probe Number = 4, T = 28.977000 C 13:39:24 Temperature Measurement, Probe Number = 5, T = 26.310000 C 13:39:24 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.45 P3 = RTD Outside, Value = 29.23 P4 = Supply Water, Value = 28.61 P5 = Return Water, Value = 28.84 P6 = @ RTD Outside, Value = 28.98 P7 = Ambient, Value = 26.31 13:39:24 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000040, Bz = -0.000020 T 13:39:32 Move Stage, position = 0.376000 13:39:33 Stage Position Measurement, pos = 0.3764950 m 13:39:34 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:39:35 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:39:35 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:39:36 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:39:36 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 13:39:37 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:39:37 Main Volt Voltage Measurement, V = 0.000002 V 13:39:38 Temperature Measurement, Probe Number = 0, T = 29.453000 C 13:39:42 Temperature Measurement, Probe Number = 1, T = 29.230000 C 13:39:42 Temperature Measurement, Probe Number = 2, T = 28.601000 C 13:39:43 Temperature Measurement, Probe Number = 3, T = 28.836000 C 13:39:43 Temperature Measurement, Probe Number = 4, T = 28.984000 C 13:39:44 Temperature Measurement, Probe Number = 5, T = 26.369000 C 13:39:44 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.45 P3 = RTD Outside, Value = 29.23 P4 = Supply Water, Value = 28.60 P5 = Return Water, Value = 28.84 P6 = @ RTD Outside, Value = 28.98 P7 = Ambient, Value = 26.37 13:39:44 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000050, Bz = -0.000010 T 13:39:52 Move Stage, position = 0.378000 13:39:53 Stage Position Measurement, pos = 0.3784850 m 13:39:54 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:39:55 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:39:55 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:39:56 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:39:56 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 13:39:57 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:39:57 Main Volt Voltage Measurement, V = 0.000002 V 13:39:58 Temperature Measurement, Probe Number = 0, T = 29.453000 C 13:40:02 Temperature Measurement, Probe Number = 1, T = 29.231000 C 13:40:02 Temperature Measurement, Probe Number = 2, T = 28.613000 C 13:40:03 Temperature Measurement, Probe Number = 3, T = 28.831000 C 13:40:03 Temperature Measurement, Probe Number = 4, T = 28.989000 C 13:40:04 Temperature Measurement, Probe Number = 5, T = 26.295000 C 13:40:04 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.45 P3 = RTD Outside, Value = 29.23 P4 = Supply Water, Value = 28.61 P5 = Return Water, Value = 28.83 P6 = @ RTD Outside, Value = 28.99 P7 = Ambient, Value = 26.30 13:40:04 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000040, Bz = -0.000020 T 13:40:12 Move Stage, position = 0.380000 13:40:13 Stage Position Measurement, pos = 0.3804950 m 13:40:14 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:40:15 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:40:16 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:40:16 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:40:16 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 13:40:17 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:40:18 Main Volt Voltage Measurement, V = 0.000002 V 13:40:18 Temperature Measurement, Probe Number = 0, T = 29.458000 C 13:40:22 Temperature Measurement, Probe Number = 1, T = 29.233000 C 13:40:22 Temperature Measurement, Probe Number = 2, T = 28.637000 C 13:40:23 Temperature Measurement, Probe Number = 3, T = 28.841000 C 13:40:24 Temperature Measurement, Probe Number = 4, T = 28.993000 C 13:40:24 Temperature Measurement, Probe Number = 5, T = 26.293000 C 13:40:24 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.46 P3 = RTD Outside, Value = 29.23 P4 = Supply Water, Value = 28.64 P5 = Return Water, Value = 28.84 P6 = @ RTD Outside, Value = 28.99 P7 = Ambient, Value = 26.29 13:40:24 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000050, Bz = -0.000020 T 13:40:32 Move Stage, position = 0.382000 13:40:33 Stage Position Measurement, pos = 0.3825000 m 13:40:34 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:40:35 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:40:36 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:40:36 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:40:36 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 13:40:37 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:40:38 Main Volt Voltage Measurement, V = 0.000002 V 13:40:38 Temperature Measurement, Probe Number = 0, T = 29.459000 C 13:40:42 Temperature Measurement, Probe Number = 1, T = 29.231000 C 13:40:42 Temperature Measurement, Probe Number = 2, T = 28.661000 C 13:40:43 Temperature Measurement, Probe Number = 3, T = 28.839000 C 13:40:44 Temperature Measurement, Probe Number = 4, T = 28.997000 C 13:40:44 Temperature Measurement, Probe Number = 5, T = 26.262000 C 13:40:44 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.46 P3 = RTD Outside, Value = 29.23 P4 = Supply Water, Value = 28.66 P5 = Return Water, Value = 28.84 P6 = @ RTD Outside, Value = 29.00 P7 = Ambient, Value = 26.26 13:40:44 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000040, Bz = -0.000020 T 13:40:52 Move Stage, position = 0.384000 13:40:54 Stage Position Measurement, pos = 0.3845150 m 13:40:54 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:40:55 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:40:56 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:40:56 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:40:56 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 13:40:57 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:40:58 Main Volt Voltage Measurement, V = 0.000002 V 13:40:58 Temperature Measurement, Probe Number = 0, T = 29.459000 C 13:41:02 Temperature Measurement, Probe Number = 1, T = 29.233000 C 13:41:03 Temperature Measurement, Probe Number = 2, T = 28.678000 C 13:41:03 Temperature Measurement, Probe Number = 3, T = 28.830000 C 13:41:04 Temperature Measurement, Probe Number = 4, T = 28.998000 C 13:41:04 Temperature Measurement, Probe Number = 5, T = 26.273000 C 13:41:04 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.46 P3 = RTD Outside, Value = 29.23 P4 = Supply Water, Value = 28.68 P5 = Return Water, Value = 28.83 P6 = @ RTD Outside, Value = 29.00 P7 = Ambient, Value = 26.27 13:41:04 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000040, Bz = -0.000020 T 13:41:12 Move Stage, position = 0.386000 13:41:14 Stage Position Measurement, pos = 0.3865550 m 13:41:14 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:41:15 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:41:16 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:41:17 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:41:17 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 13:41:17 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:41:18 Main Volt Voltage Measurement, V = 0.000002 V 13:41:18 Temperature Measurement, Probe Number = 0, T = 29.456000 C 13:41:22 Temperature Measurement, Probe Number = 1, T = 29.221000 C 13:41:23 Temperature Measurement, Probe Number = 2, T = 28.660000 C 13:41:23 Temperature Measurement, Probe Number = 3, T = 28.829000 C 13:41:24 Temperature Measurement, Probe Number = 4, T = 28.993000 C 13:41:24 Temperature Measurement, Probe Number = 5, T = 26.288000 C 13:41:24 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.46 P3 = RTD Outside, Value = 29.22 P4 = Supply Water, Value = 28.66 P5 = Return Water, Value = 28.83 P6 = @ RTD Outside, Value = 28.99 P7 = Ambient, Value = 26.29 13:41:24 3 Axis Hall Probe Field Measurement, Bx = -0.000010, By = 0.000050, Bz = -0.000020 T 13:41:32 Move Stage, position = 0.388000 13:41:34 Stage Position Measurement, pos = 0.3885300 m 13:41:35 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:41:35 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:41:36 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:41:37 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:41:37 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 13:41:37 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:41:38 Main Volt Voltage Measurement, V = 0.000002 V 13:41:39 Temperature Measurement, Probe Number = 0, T = 29.455000 C 13:41:42 Temperature Measurement, Probe Number = 1, T = 29.223000 C 13:41:43 Temperature Measurement, Probe Number = 2, T = 28.631000 C 13:41:43 Temperature Measurement, Probe Number = 3, T = 28.843000 C 13:41:44 Temperature Measurement, Probe Number = 4, T = 28.986000 C 13:41:45 Temperature Measurement, Probe Number = 5, T = 26.302000 C 13:41:45 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.45 P3 = RTD Outside, Value = 29.22 P4 = Supply Water, Value = 28.63 P5 = Return Water, Value = 28.84 P6 = @ RTD Outside, Value = 28.99 P7 = Ambient, Value = 26.30 13:41:45 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000050, Bz = -0.000010 T 13:41:52 Move Stage, position = 0.390000 13:41:54 Stage Position Measurement, pos = 0.3905300 m 13:41:55 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:41:55 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:41:56 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:41:57 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:41:57 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 13:41:57 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:41:58 Main Volt Voltage Measurement, V = 0.000002 V 13:41:59 Temperature Measurement, Probe Number = 0, T = 29.455000 C 13:42:02 Temperature Measurement, Probe Number = 1, T = 29.230000 C 13:42:03 Temperature Measurement, Probe Number = 2, T = 28.610000 C 13:42:03 Temperature Measurement, Probe Number = 3, T = 28.847000 C 13:42:04 Temperature Measurement, Probe Number = 4, T = 28.989000 C 13:42:05 Temperature Measurement, Probe Number = 5, T = 26.355000 C 13:42:05 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.45 P3 = RTD Outside, Value = 29.23 P4 = Supply Water, Value = 28.61 P5 = Return Water, Value = 28.85 P6 = @ RTD Outside, Value = 28.99 P7 = Ambient, Value = 26.36 13:42:05 3 Axis Hall Probe Field Measurement, Bx = -0.000010, By = 0.000040, Bz = -0.000010 T 13:42:12 Move Stage, position = 0.392000 13:42:14 Stage Position Measurement, pos = 0.3924950 m 13:42:15 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:42:15 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:42:16 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:42:17 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:42:17 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 13:42:18 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:42:18 Main Volt Voltage Measurement, V = 0.000002 V 13:42:19 Temperature Measurement, Probe Number = 0, T = 29.456000 C 13:42:22 Temperature Measurement, Probe Number = 1, T = 29.233000 C 13:42:23 Temperature Measurement, Probe Number = 2, T = 28.617000 C 13:42:24 Temperature Measurement, Probe Number = 3, T = 28.856000 C 13:42:24 Temperature Measurement, Probe Number = 4, T = 28.993000 C 13:42:25 Temperature Measurement, Probe Number = 5, T = 26.339000 C 13:42:25 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.46 P3 = RTD Outside, Value = 29.23 P4 = Supply Water, Value = 28.62 P5 = Return Water, Value = 28.86 P6 = @ RTD Outside, Value = 28.99 P7 = Ambient, Value = 26.34 13:42:25 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000030, Bz = -0.000030 T 13:42:33 Move Stage, position = 0.394000 13:42:34 Stage Position Measurement, pos = 0.3945250 m 13:42:35 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:42:36 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:42:36 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:42:37 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:42:37 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 13:42:38 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:42:38 Main Volt Voltage Measurement, V = 0.000002 V 13:42:39 Temperature Measurement, Probe Number = 0, T = 29.457000 C 13:42:43 Temperature Measurement, Probe Number = 1, T = 29.235000 C 13:42:43 Temperature Measurement, Probe Number = 2, T = 28.645000 C 13:42:44 Temperature Measurement, Probe Number = 3, T = 28.857000 C 13:42:44 Temperature Measurement, Probe Number = 4, T = 28.997000 C 13:42:45 Temperature Measurement, Probe Number = 5, T = 26.377000 C 13:42:45 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.46 P3 = RTD Outside, Value = 29.23 P4 = Supply Water, Value = 28.64 P5 = Return Water, Value = 28.86 P6 = @ RTD Outside, Value = 29.00 P7 = Ambient, Value = 26.38 13:42:45 3 Axis Hall Probe Field Measurement, Bx = -0.000010, By = 0.000040, Bz = -0.000020 T 13:42:53 Move Stage, position = 0.396000 13:42:54 Stage Position Measurement, pos = 0.3965200 m 13:42:55 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:42:56 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:42:56 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:42:57 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:42:57 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 13:42:58 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:42:59 Main Volt Voltage Measurement, V = 0.000002 V 13:42:59 Temperature Measurement, Probe Number = 0, T = 29.458000 C 13:43:03 Temperature Measurement, Probe Number = 1, T = 29.237000 C 13:43:03 Temperature Measurement, Probe Number = 2, T = 28.663000 C 13:43:04 Temperature Measurement, Probe Number = 3, T = 28.847000 C 13:43:05 Temperature Measurement, Probe Number = 4, T = 29.003000 C 13:43:05 Temperature Measurement, Probe Number = 5, T = 26.291000 C 13:43:05 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.46 P3 = RTD Outside, Value = 29.24 P4 = Supply Water, Value = 28.66 P5 = Return Water, Value = 28.85 P6 = @ RTD Outside, Value = 29.00 P7 = Ambient, Value = 26.29 13:43:05 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000040, Bz = -0.000030 T 13:43:13 Move Stage, position = 0.398000 13:43:15 Stage Position Measurement, pos = 0.3985350 m 13:43:15 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:43:16 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:43:17 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:43:17 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:43:17 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 13:43:18 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:43:19 Main Volt Voltage Measurement, V = 0.000002 V 13:43:19 Temperature Measurement, Probe Number = 0, T = 29.459000 C 13:43:23 Temperature Measurement, Probe Number = 1, T = 29.237000 C 13:43:23 Temperature Measurement, Probe Number = 2, T = 28.676000 C 13:43:24 Temperature Measurement, Probe Number = 3, T = 28.850000 C 13:43:25 Temperature Measurement, Probe Number = 4, T = 29.008000 C 13:43:25 Temperature Measurement, Probe Number = 5, T = 26.281000 C 13:43:25 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.46 P3 = RTD Outside, Value = 29.24 P4 = Supply Water, Value = 28.68 P5 = Return Water, Value = 28.85 P6 = @ RTD Outside, Value = 29.01 P7 = Ambient, Value = 26.28 13:43:25 3 Axis Hall Probe Field Measurement, Bx = -0.000010, By = 0.000040, Bz = -0.000020 T 13:43:33 Move Stage, position = 0.400000 13:43:35 Stage Position Measurement, pos = 0.4005450 m 13:43:35 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:43:36 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:43:37 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:43:37 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:43:37 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 13:43:38 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:43:39 Main Volt Voltage Measurement, V = 0.000002 V 13:43:39 Temperature Measurement, Probe Number = 0, T = 29.459000 C 13:43:43 Temperature Measurement, Probe Number = 1, T = 29.237000 C 13:43:44 Temperature Measurement, Probe Number = 2, T = 28.694000 C 13:43:44 Temperature Measurement, Probe Number = 3, T = 28.865000 C 13:43:45 Temperature Measurement, Probe Number = 4, T = 29.009000 C 13:43:45 Temperature Measurement, Probe Number = 5, T = 26.272000 C 13:43:45 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.46 P3 = RTD Outside, Value = 29.24 P4 = Supply Water, Value = 28.69 P5 = Return Water, Value = 28.86 P6 = @ RTD Outside, Value = 29.01 P7 = Ambient, Value = 26.27 13:43:45 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000050, Bz = -0.000020 T 13:43:53 Move Stage, position = 0.402000 13:43:55 Stage Position Measurement, pos = 0.4025550 m 13:43:55 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:43:56 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:43:57 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:43:58 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:43:58 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 13:43:58 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:43:59 Main Volt Voltage Measurement, V = 0.000002 V 13:44:00 Temperature Measurement, Probe Number = 0, T = 29.458000 C 13:44:03 Temperature Measurement, Probe Number = 1, T = 29.235000 C 13:44:04 Temperature Measurement, Probe Number = 2, T = 28.702000 C 13:44:04 Temperature Measurement, Probe Number = 3, T = 28.878000 C 13:44:05 Temperature Measurement, Probe Number = 4, T = 29.011000 C 13:44:05 Temperature Measurement, Probe Number = 5, T = 26.315000 C 13:44:05 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.46 P3 = RTD Outside, Value = 29.23 P4 = Supply Water, Value = 28.70 P5 = Return Water, Value = 28.88 P6 = @ RTD Outside, Value = 29.01 P7 = Ambient, Value = 26.32 13:44:06 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000040, Bz = -0.000020 T 13:44:13 Move Stage, position = 0.404000 13:44:15 Stage Position Measurement, pos = 0.4045650 m 13:44:16 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:44:16 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:44:17 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:44:18 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:44:18 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 13:44:18 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:44:19 Main Volt Voltage Measurement, V = 0.000002 V 13:44:20 Temperature Measurement, Probe Number = 0, T = 29.455000 C 13:44:23 Temperature Measurement, Probe Number = 1, T = 29.230000 C 13:44:24 Temperature Measurement, Probe Number = 2, T = 28.709000 C 13:44:24 Temperature Measurement, Probe Number = 3, T = 28.871000 C 13:44:25 Temperature Measurement, Probe Number = 4, T = 29.009000 C 13:44:26 Temperature Measurement, Probe Number = 5, T = 26.281000 C 13:44:26 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.45 P3 = RTD Outside, Value = 29.23 P4 = Supply Water, Value = 28.71 P5 = Return Water, Value = 28.87 P6 = @ RTD Outside, Value = 29.01 P7 = Ambient, Value = 26.28 13:44:26 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000040, Bz = -0.000020 T 13:44:33 Move Stage, position = 0.406000 13:44:35 Stage Position Measurement, pos = 0.4065500 m 13:44:36 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:44:36 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:44:37 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:44:38 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:44:38 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 13:44:39 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:44:39 Main Volt Voltage Measurement, V = 0.000002 V 13:44:40 Temperature Measurement, Probe Number = 0, T = 29.452000 C 13:44:43 Temperature Measurement, Probe Number = 1, T = 29.232000 C 13:44:44 Temperature Measurement, Probe Number = 2, T = 28.697000 C 13:44:45 Temperature Measurement, Probe Number = 3, T = 28.866000 C 13:44:45 Temperature Measurement, Probe Number = 4, T = 29.010000 C 13:44:46 Temperature Measurement, Probe Number = 5, T = 26.241000 C 13:44:46 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.45 P3 = RTD Outside, Value = 29.23 P4 = Supply Water, Value = 28.70 P5 = Return Water, Value = 28.87 P6 = @ RTD Outside, Value = 29.01 P7 = Ambient, Value = 26.24 13:44:46 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000040, Bz = -0.000020 T 13:44:53 Move Stage, position = 0.408000 13:44:55 Stage Position Measurement, pos = 0.4085450 m 13:44:56 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:44:57 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:44:57 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:44:58 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:44:58 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 13:44:59 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:44:59 Main Volt Voltage Measurement, V = 0.000002 V 13:45:00 Temperature Measurement, Probe Number = 0, T = 29.446000 C 13:45:03 Temperature Measurement, Probe Number = 1, T = 29.227000 C 13:45:04 Temperature Measurement, Probe Number = 2, T = 28.652000 C 13:45:05 Temperature Measurement, Probe Number = 3, T = 28.822000 C 13:45:05 Temperature Measurement, Probe Number = 4, T = 29.009000 C 13:45:06 Temperature Measurement, Probe Number = 5, T = 26.194000 C 13:45:06 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.45 P3 = RTD Outside, Value = 29.23 P4 = Supply Water, Value = 28.65 P5 = Return Water, Value = 28.82 P6 = @ RTD Outside, Value = 29.01 P7 = Ambient, Value = 26.19 13:45:06 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000040, Bz = -0.000030 T 13:45:14 Move Stage, position = 0.410000 13:45:15 Stage Position Measurement, pos = 0.4105200 m 13:45:16 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:45:17 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:45:17 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:45:18 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:45:18 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 13:45:19 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:45:19 Main Volt Voltage Measurement, V = 0.000003 V 13:45:20 Temperature Measurement, Probe Number = 0, T = 29.443000 C 13:45:24 Temperature Measurement, Probe Number = 1, T = 29.222000 C 13:45:24 Temperature Measurement, Probe Number = 2, T = 28.611000 C 13:45:25 Temperature Measurement, Probe Number = 3, T = 28.776000 C 13:45:25 Temperature Measurement, Probe Number = 4, T = 29.005000 C 13:45:26 Temperature Measurement, Probe Number = 5, T = 26.171000 C 13:45:26 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.44 P3 = RTD Outside, Value = 29.22 P4 = Supply Water, Value = 28.61 P5 = Return Water, Value = 28.78 P6 = @ RTD Outside, Value = 29.00 P7 = Ambient, Value = 26.17 13:45:26 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000040, Bz = -0.000020 T 13:45:34 Move Stage, position = 0.412000 13:45:35 Stage Position Measurement, pos = 0.4125250 m 13:45:36 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:45:37 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:45:38 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:45:38 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:45:38 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 13:45:39 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:45:40 Main Volt Voltage Measurement, V = 0.000002 V 13:45:40 Temperature Measurement, Probe Number = 0, T = 29.437000 C 13:45:44 Temperature Measurement, Probe Number = 1, T = 29.220000 C 13:45:44 Temperature Measurement, Probe Number = 2, T = 28.594000 C 13:45:45 Temperature Measurement, Probe Number = 3, T = 28.738000 C 13:45:46 Temperature Measurement, Probe Number = 4, T = 29.003000 C 13:45:46 Temperature Measurement, Probe Number = 5, T = 26.109000 C 13:45:46 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.44 P3 = RTD Outside, Value = 29.22 P4 = Supply Water, Value = 28.59 P5 = Return Water, Value = 28.74 P6 = @ RTD Outside, Value = 29.00 P7 = Ambient, Value = 26.11 13:45:46 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000040, Bz = -0.000030 T 13:45:54 Move Stage, position = 0.414000 13:45:56 Stage Position Measurement, pos = 0.4145200 m 13:45:56 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:45:57 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:45:58 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:45:58 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:45:58 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 13:45:59 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:46:00 Main Volt Voltage Measurement, V = 0.000003 V 13:46:00 Temperature Measurement, Probe Number = 0, T = 29.433000 C 13:46:04 Temperature Measurement, Probe Number = 1, T = 29.212000 C 13:46:05 Temperature Measurement, Probe Number = 2, T = 28.587000 C 13:46:05 Temperature Measurement, Probe Number = 3, T = 28.730000 C 13:46:06 Temperature Measurement, Probe Number = 4, T = 28.997000 C 13:46:06 Temperature Measurement, Probe Number = 5, T = 26.160000 C 13:46:06 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.43 P3 = RTD Outside, Value = 29.21 P4 = Supply Water, Value = 28.59 P5 = Return Water, Value = 28.73 P6 = @ RTD Outside, Value = 29.00 P7 = Ambient, Value = 26.16 13:46:06 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000040, Bz = -0.000010 T 13:46:14 Move Stage, position = 0.416000 13:46:16 Stage Position Measurement, pos = 0.4165350 m 13:46:16 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:46:17 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:46:18 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:46:19 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:46:19 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 13:46:19 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:46:20 Main Volt Voltage Measurement, V = 0.000003 V 13:46:21 Temperature Measurement, Probe Number = 0, T = 29.431000 C 13:46:24 Temperature Measurement, Probe Number = 1, T = 29.213000 C 13:46:25 Temperature Measurement, Probe Number = 2, T = 28.586000 C 13:46:25 Temperature Measurement, Probe Number = 3, T = 28.746000 C 13:46:26 Temperature Measurement, Probe Number = 4, T = 28.994000 C 13:46:26 Temperature Measurement, Probe Number = 5, T = 26.166000 C 13:46:26 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.43 P3 = RTD Outside, Value = 29.21 P4 = Supply Water, Value = 28.59 P5 = Return Water, Value = 28.75 P6 = @ RTD Outside, Value = 28.99 P7 = Ambient, Value = 26.17 13:46:27 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000040, Bz = -0.000020 T 13:46:34 Move Stage, position = 0.418000 13:46:36 Stage Position Measurement, pos = 0.4185450 m 13:46:37 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:46:37 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:46:38 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:46:39 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:46:39 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 13:46:39 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:46:40 Main Volt Voltage Measurement, V = 0.000002 V 13:46:41 Temperature Measurement, Probe Number = 0, T = 29.429000 C 13:46:44 Temperature Measurement, Probe Number = 1, T = 29.202000 C 13:46:45 Temperature Measurement, Probe Number = 2, T = 28.541000 C 13:46:45 Temperature Measurement, Probe Number = 3, T = 28.757000 C 13:46:46 Temperature Measurement, Probe Number = 4, T = 28.975000 C 13:46:47 Temperature Measurement, Probe Number = 5, T = 26.137000 C 13:46:47 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.43 P3 = RTD Outside, Value = 29.20 P4 = Supply Water, Value = 28.54 P5 = Return Water, Value = 28.76 P6 = @ RTD Outside, Value = 28.98 P7 = Ambient, Value = 26.14 13:46:47 3 Axis Hall Probe Field Measurement, Bx = -0.000010, By = 0.000050, Bz = -0.000010 T 13:46:54 Move Stage, position = 0.420000 13:46:56 Stage Position Measurement, pos = 0.4205550 m 13:46:57 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:46:57 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:46:58 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:46:59 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:46:59 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 13:47:00 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:47:00 Main Volt Voltage Measurement, V = 0.000002 V 13:47:01 Temperature Measurement, Probe Number = 0, T = 29.428000 C 13:47:04 Temperature Measurement, Probe Number = 1, T = 29.189000 C 13:47:05 Temperature Measurement, Probe Number = 2, T = 28.498000 C 13:47:06 Temperature Measurement, Probe Number = 3, T = 28.764000 C 13:47:06 Temperature Measurement, Probe Number = 4, T = 28.951000 C 13:47:07 Temperature Measurement, Probe Number = 5, T = 26.312000 C 13:47:07 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.43 P3 = RTD Outside, Value = 29.19 P4 = Supply Water, Value = 28.50 P5 = Return Water, Value = 28.76 P6 = @ RTD Outside, Value = 28.95 P7 = Ambient, Value = 26.31 13:47:07 3 Axis Hall Probe Field Measurement, Bx = -0.000010, By = 0.000050, Bz = -0.000020 T 13:47:14 Move Stage, position = 0.422000 13:47:16 Stage Position Measurement, pos = 0.4225450 m 13:47:17 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:47:18 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:47:18 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:47:19 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:47:19 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 13:47:20 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:47:20 Main Volt Voltage Measurement, V = 0.000002 V 13:47:21 Temperature Measurement, Probe Number = 0, T = 29.426000 C 13:47:24 Temperature Measurement, Probe Number = 1, T = 29.189000 C 13:47:25 Temperature Measurement, Probe Number = 2, T = 28.509000 C 13:47:26 Temperature Measurement, Probe Number = 3, T = 28.763000 C 13:47:26 Temperature Measurement, Probe Number = 4, T = 28.948000 C 13:47:27 Temperature Measurement, Probe Number = 5, T = 26.326000 C 13:47:27 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.43 P3 = RTD Outside, Value = 29.19 P4 = Supply Water, Value = 28.51 P5 = Return Water, Value = 28.76 P6 = @ RTD Outside, Value = 28.95 P7 = Ambient, Value = 26.33 13:47:27 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000040, Bz = -0.000020 T 13:47:35 Move Stage, position = 0.424000 13:47:36 Stage Position Measurement, pos = 0.4245250 m 13:47:37 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:47:38 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:47:38 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:47:39 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:47:39 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 13:47:40 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:47:40 Main Volt Voltage Measurement, V = 0.000002 V 13:47:41 Temperature Measurement, Probe Number = 0, T = 29.428000 C 13:47:45 Temperature Measurement, Probe Number = 1, T = 29.197000 C 13:47:45 Temperature Measurement, Probe Number = 2, T = 28.496000 C 13:47:46 Temperature Measurement, Probe Number = 3, T = 28.764000 C 13:47:46 Temperature Measurement, Probe Number = 4, T = 28.952000 C 13:47:47 Temperature Measurement, Probe Number = 5, T = 26.288000 C 13:47:47 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.43 P3 = RTD Outside, Value = 29.20 P4 = Supply Water, Value = 28.50 P5 = Return Water, Value = 28.76 P6 = @ RTD Outside, Value = 28.95 P7 = Ambient, Value = 26.29 13:47:47 3 Axis Hall Probe Field Measurement, Bx = -0.000010, By = 0.000050, Bz = -0.000030 T 13:47:55 Move Stage, position = 0.426000 13:47:56 Stage Position Measurement, pos = 0.4265400 m 13:47:57 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:47:58 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:47:58 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:47:59 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:47:59 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 13:48:00 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:48:00 Main Volt Voltage Measurement, V = 0.000002 V 13:48:01 Temperature Measurement, Probe Number = 0, T = 29.427000 C 13:48:05 Temperature Measurement, Probe Number = 1, T = 29.199000 C 13:48:05 Temperature Measurement, Probe Number = 2, T = 28.494000 C 13:48:06 Temperature Measurement, Probe Number = 3, T = 28.751000 C 13:48:06 Temperature Measurement, Probe Number = 4, T = 28.959000 C 13:48:07 Temperature Measurement, Probe Number = 5, T = 26.208000 C 13:48:07 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.43 P3 = RTD Outside, Value = 29.20 P4 = Supply Water, Value = 28.49 P5 = Return Water, Value = 28.75 P6 = @ RTD Outside, Value = 28.96 P7 = Ambient, Value = 26.21 13:48:07 3 Axis Hall Probe Field Measurement, Bx = -0.000010, By = 0.000050, Bz = -0.000010 T 13:48:15 Move Stage, position = 0.428000 13:48:16 Stage Position Measurement, pos = 0.4285200 m 13:48:17 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:48:18 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:48:19 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:48:19 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:48:19 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 13:48:20 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:48:21 Main Volt Voltage Measurement, V = 0.000002 V 13:48:21 Temperature Measurement, Probe Number = 0, T = 29.426000 C 13:48:25 Temperature Measurement, Probe Number = 1, T = 29.204000 C 13:48:25 Temperature Measurement, Probe Number = 2, T = 28.479000 C 13:48:26 Temperature Measurement, Probe Number = 3, T = 28.740000 C 13:48:27 Temperature Measurement, Probe Number = 4, T = 28.967000 C 13:48:27 Temperature Measurement, Probe Number = 5, T = 26.280000 C 13:48:27 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.43 P3 = RTD Outside, Value = 29.20 P4 = Supply Water, Value = 28.48 P5 = Return Water, Value = 28.74 P6 = @ RTD Outside, Value = 28.97 P7 = Ambient, Value = 26.28 13:48:27 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000040, Bz = -0.000020 T 13:48:35 Move Stage, position = 0.430000 13:48:37 Stage Position Measurement, pos = 0.4305250 m 13:48:37 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:48:38 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:48:39 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:48:39 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:48:39 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 13:48:40 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:48:41 Main Volt Voltage Measurement, V = 0.000002 V 13:48:41 Temperature Measurement, Probe Number = 0, T = 29.425000 C 13:48:45 Temperature Measurement, Probe Number = 1, T = 29.203000 C 13:48:45 Temperature Measurement, Probe Number = 2, T = 28.491000 C 13:48:46 Temperature Measurement, Probe Number = 3, T = 28.724000 C 13:48:47 Temperature Measurement, Probe Number = 4, T = 28.972000 C 13:48:47 Temperature Measurement, Probe Number = 5, T = 26.254000 C 13:48:47 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.43 P3 = RTD Outside, Value = 29.20 P4 = Supply Water, Value = 28.49 P5 = Return Water, Value = 28.72 P6 = @ RTD Outside, Value = 28.97 P7 = Ambient, Value = 26.25 13:48:47 3 Axis Hall Probe Field Measurement, Bx = -0.000010, By = 0.000040, Bz = -0.000020 T 13:48:55 Move Stage, position = 0.432000 13:48:57 Stage Position Measurement, pos = 0.4325000 m 13:48:57 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:48:58 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:48:59 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:49:00 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:49:00 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 13:49:00 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:49:01 Main Volt Voltage Measurement, V = 0.000003 V 13:49:01 Temperature Measurement, Probe Number = 0, T = 29.422000 C 13:49:05 Temperature Measurement, Probe Number = 1, T = 29.203000 C 13:49:06 Temperature Measurement, Probe Number = 2, T = 28.502000 C 13:49:06 Temperature Measurement, Probe Number = 3, T = 28.677000 C 13:49:07 Temperature Measurement, Probe Number = 4, T = 28.973000 C 13:49:07 Temperature Measurement, Probe Number = 5, T = 26.219000 C 13:49:07 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.42 P3 = RTD Outside, Value = 29.20 P4 = Supply Water, Value = 28.50 P5 = Return Water, Value = 28.68 P6 = @ RTD Outside, Value = 28.97 P7 = Ambient, Value = 26.22 13:49:07 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000040, Bz = -0.000020 T 13:49:15 Move Stage, position = 0.434000 13:49:17 Stage Position Measurement, pos = 0.4345200 m 13:49:17 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:49:18 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:49:19 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:49:20 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:49:20 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 13:49:20 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:49:21 Main Volt Voltage Measurement, V = 0.000002 V 13:49:22 Temperature Measurement, Probe Number = 0, T = 29.421000 C 13:49:25 Temperature Measurement, Probe Number = 1, T = 29.203000 C 13:49:26 Temperature Measurement, Probe Number = 2, T = 28.518000 C 13:49:26 Temperature Measurement, Probe Number = 3, T = 28.693000 C 13:49:27 Temperature Measurement, Probe Number = 4, T = 28.976000 C 13:49:27 Temperature Measurement, Probe Number = 5, T = 26.193000 C 13:49:27 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.42 P3 = RTD Outside, Value = 29.20 P4 = Supply Water, Value = 28.52 P5 = Return Water, Value = 28.69 P6 = @ RTD Outside, Value = 28.98 P7 = Ambient, Value = 26.19 13:49:28 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000050, Bz = -0.000020 T 13:49:35 Move Stage, position = 0.436000 13:49:37 Stage Position Measurement, pos = 0.4365450 m 13:49:38 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:49:38 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:49:39 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:49:40 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:49:40 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 13:49:40 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:49:41 Main Volt Voltage Measurement, V = 0.000002 V 13:49:42 Temperature Measurement, Probe Number = 0, T = 29.420000 C 13:49:45 Temperature Measurement, Probe Number = 1, T = 29.202000 C 13:49:46 Temperature Measurement, Probe Number = 2, T = 28.539000 C 13:49:46 Temperature Measurement, Probe Number = 3, T = 28.730000 C 13:49:47 Temperature Measurement, Probe Number = 4, T = 28.978000 C 13:49:48 Temperature Measurement, Probe Number = 5, T = 26.245000 C 13:49:48 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.42 P3 = RTD Outside, Value = 29.20 P4 = Supply Water, Value = 28.54 P5 = Return Water, Value = 28.73 P6 = @ RTD Outside, Value = 28.98 P7 = Ambient, Value = 26.25 13:49:48 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000040, Bz = -0.000020 T 13:49:55 Move Stage, position = 0.438000 13:49:57 Stage Position Measurement, pos = 0.4385550 m 13:49:58 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:49:58 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:49:59 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:50:00 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:50:00 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 13:50:01 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:50:01 Main Volt Voltage Measurement, V = 0.000002 V 13:50:02 Temperature Measurement, Probe Number = 0, T = 29.418000 C 13:50:05 Temperature Measurement, Probe Number = 1, T = 29.200000 C 13:50:06 Temperature Measurement, Probe Number = 2, T = 28.519000 C 13:50:07 Temperature Measurement, Probe Number = 3, T = 28.753000 C 13:50:07 Temperature Measurement, Probe Number = 4, T = 28.978000 C 13:50:08 Temperature Measurement, Probe Number = 5, T = 26.219000 C 13:50:08 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.42 P3 = RTD Outside, Value = 29.20 P4 = Supply Water, Value = 28.52 P5 = Return Water, Value = 28.75 P6 = @ RTD Outside, Value = 28.98 P7 = Ambient, Value = 26.22 13:50:08 3 Axis Hall Probe Field Measurement, Bx = 0.000000, By = 0.000050, Bz = -0.000020 T 13:50:15 Move Stage, position = 0.440000 13:50:17 Stage Position Measurement, pos = 0.4405500 m 13:50:18 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:50:19 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:50:19 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:50:20 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000058 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:50:20 Magnet Avgerage Current Measurement for Main : Imag = 0.0000 +- 0.0000 A 13:50:21 Magnet Current for Main, Imag = 0.0000 A Vtrans = -0.000057 V, Factor V/A = 0.200000 V/A, Zero Current = -0.000294 A 13:50:21 Main Volt Voltage Measurement, V = 0.000003 V 13:50:22 Temperature Measurement, Probe Number = 0, T = 29.418000 C 13:50:25 Temperature Measurement, Probe Number = 1, T = 29.197000 C 13:50:26 Temperature Measurement, Probe Number = 2, T = 28.503000 C 13:50:27 Temperature Measurement, Probe Number = 3, T = 28.775000 C 13:50:27 Temperature Measurement, Probe Number = 4, T = 28.973000 C 13:50:28 Temperature Measurement, Probe Number = 5, T = 26.219000 C 13:50:28 Magnet Test Parameter Measurements P0 = Main, Value = 0.00 P1 = Main Volt, Value = 0.00 P2 = RTD Inside, Value = 29.42 P3 = RTD Outside, Value = 29.20 P4 = Supply Water, Value = 28.50 P5 = Return Water, Value = 28.77 P6 = @ RTD Outside, Value = 28.97 P7 = Ambient, Value = 26.22 13:50:28 3 Axis Hall Probe Field Measurement, Bx = 0.000010, By = 0.000040, Bz = -0.000020 T 13:51:34 Move Stage, position = 0.000000 13:51:35 Magnet system Main current ramping to 0.0000 A...