Wire detector tooling ball 1 offset calibrations values are below: Detector 1 TB1 to Wire Detector offset 47.5955 mm Detector 2 TB1 to Wire Detector offset 47.3295 mm Updated 12/11/2017 because laser was replaced, was 47.2799. Difference is 49.6 microns Detector 3 TB1 to Wire Detector offset 47.4922 mm Detector 4 TB1 to Wire Detector offset 47.3683 mm The detectors were refurbished with new slits and tensioning springs, so the tooling ball 1 to detector offsets had to be redone. The measurement data for each detector is in file Flip Test Analysis for Detector N Final.txt where N is the detector number. Detectors 1,2 and 4 calibrated easily and their values are from runs 1-7. After run 1-7 it was clear detector 2 needed screw tightening so its values come from runs 8-11, which are after the screw tightening. During runs 8-11 only detector 2 was flipped, so detectors 1, 2 and 4 data for those runs are no good. Detector 2 was recalibrated on 12/11/2017 after laser failed and then was replaced. Offset was 47.2799. Difference is 49.6 microns.