SLAC Magnetic Measurements Date: 01-07-2021 Time: 12:06:39 Project: LCLS-II Magnet Type: Radiation_Detection_System Magnet Name: SXU_017_Cell_31 Serial Number: PC33408 Run Number: 15 Measurement Device: Coil Operator: sda Comment: Coil Integral measurement in tunnel. Full Set 15 12:06:57 Ambient Measurement Device(s) Being Used, type = None 12:06:57 Magnet Voltage Measurement Device(s) Being Used, type = None 12:06:57 Temperature Measurement Device(s) Being Used, type = HPTHERM_HP34970 12:06:57 XMAG Device(s) Being Used, type = None 12:06:57 Ambient Measurement Device(s) Being Used, type = None 12:07:03   Voltage Measurement, V = 0.000000 V 12:07:04 Temperature Measurement, Probe Number = 0, T = 20.668000 C 12:07:05 Temperature Measurement, Probe Number = 1, T = 21.515000 C 12:07:05 Magnet Test Parameter Measurements P0 = Main (A) , Value = 0.00 P1 = Walker_Meas_Junction, Value = 20.67 P2 = 3458_Meas_Junction, Value = 21.52 12:07:05 Integrated Voltage Measurement... 12:07:32 Voltage Offset Correction: v_off_ini = 0.000000817 V, v_off_fin = 0.000000704 V 12:07:32 Integrated Voltage Measurement Delta Gap = 0.0040001 m VT = 0.142309721 Vs 12:07:32 Integrated Voltage Measurement... 12:08:05 Voltage Offset Correction: v_off_ini = 0.000000835 V, v_off_fin = -0.000003031 V 12:08:06 Integrated Voltage Measurement Delta Gap = -0.0040008 m VT = -0.142348762 Vs 12:08:07   Voltage Measurement, V = 0.000000 V 12:08:07 Temperature Measurement, Probe Number = 0, T = 20.659000 C 12:08:08 Temperature Measurement, Probe Number = 1, T = 21.518000 C 12:08:08 Magnet Test Parameter Measurements P0 = Main (A) , Value = 0.00 P1 = Walker_Meas_Junction, Value = 20.66 P2 = 3458_Meas_Junction, Value = 21.52 12:08:08 Integrated Voltage Measurement... 12:08:38 Voltage Offset Correction: v_off_ini = -0.000000717 V, v_off_fin = 0.000001645 V 12:08:38 Integrated Voltage Measurement Delta Gap = 0.0040014 m VT = 0.142397229 Vs 12:08:39 Integrated Voltage Measurement... 12:09:09 Voltage Offset Correction: v_off_ini = 0.000000729 V, v_off_fin = -0.000002621 V 12:09:09 Integrated Voltage Measurement Delta Gap = -0.0040008 m VT = -0.142345018 Vs 12:09:10   Voltage Measurement, V = 0.000000 V 12:09:11 Temperature Measurement, Probe Number = 0, T = 20.645000 C 12:09:11 Temperature Measurement, Probe Number = 1, T = 21.520000 C 12:09:11 Magnet Test Parameter Measurements P0 = Main (A) , Value = 0.00 P1 = Walker_Meas_Junction, Value = 20.64 P2 = 3458_Meas_Junction, Value = 21.52 12:09:11 Integrated Voltage Measurement... 12:09:41 Voltage Offset Correction: v_off_ini = -0.000000937 V, v_off_fin = 0.000000686 V 12:09:42 Integrated Voltage Measurement Delta Gap = 0.0040011 m VT = 0.142393711 Vs 12:09:42 Integrated Voltage Measurement... 12:10:11 Voltage Offset Correction: v_off_ini = 0.000001044 V, v_off_fin = -0.000001531 V 12:10:11 Integrated Voltage Measurement Delta Gap = -0.0040007 m VT = -0.142341743 Vs 12:10:12   Voltage Measurement, V = 0.000000 V 12:10:13 Temperature Measurement, Probe Number = 0, T = 20.628000 C 12:10:14 Temperature Measurement, Probe Number = 1, T = 21.539000 C 12:10:14 Magnet Test Parameter Measurements P0 = Main (A) , Value = 0.00 P1 = Walker_Meas_Junction, Value = 20.63 P2 = 3458_Meas_Junction, Value = 21.54 12:10:14 Integrated Voltage Measurement... 12:10:44 Voltage Offset Correction: v_off_ini = -0.000001033 V, v_off_fin = 0.000001016 V 12:10:44 Integrated Voltage Measurement Delta Gap = 0.0040012 m VT = 0.142385820 Vs 12:10:44 Integrated Voltage Measurement... 12:11:15 Voltage Offset Correction: v_off_ini = 0.000000573 V, v_off_fin = -0.000003242 V 12:11:15 Integrated Voltage Measurement Delta Gap = -0.0040009 m VT = -0.142344541 Vs 12:11:16   Voltage Measurement, V = 0.000000 V 12:11:16 Temperature Measurement, Probe Number = 0, T = 20.616000 C 12:11:17 Temperature Measurement, Probe Number = 1, T = 21.530000 C 12:11:17 Magnet Test Parameter Measurements P0 = Main (A) , Value = 0.00 P1 = Walker_Meas_Junction, Value = 20.62 P2 = 3458_Meas_Junction, Value = 21.53 12:11:17 Integrated Voltage Measurement... 12:11:55 Voltage Offset Correction: v_off_ini = -0.000000934 V, v_off_fin = 0.000000938 V 12:11:55 Integrated Voltage Measurement Delta Gap = 0.0040013 m VT = 0.142393737 Vs 12:11:55 Integrated Voltage Measurement... 12:12:32 Voltage Offset Correction: v_off_ini = 0.000000518 V, v_off_fin = -0.000003596 V 12:12:32 Integrated Voltage Measurement Delta Gap = -0.0040006 m VT = -0.142333496 Vs 12:12:33   Voltage Measurement, V = 0.000000 V 12:12:34 Temperature Measurement, Probe Number = 0, T = 20.599000 C 12:12:34 Temperature Measurement, Probe Number = 1, T = 21.537000 C 12:12:34 Magnet Test Parameter Measurements P0 = Main (A) , Value = 0.00 P1 = Walker_Meas_Junction, Value = 20.60 P2 = 3458_Meas_Junction, Value = 21.54 12:12:34 Integrated Voltage Measurement... 12:13:05 Voltage Offset Correction: v_off_ini = -0.000001068 V, v_off_fin = 0.000001670 V 12:13:05 Integrated Voltage Measurement Delta Gap = 0.0040016 m VT = 0.142407854 Vs 12:13:05 Integrated Voltage Measurement... 12:13:39 Voltage Offset Correction: v_off_ini = 0.000000725 V, v_off_fin = -0.000001386 V 12:13:39 Integrated Voltage Measurement Delta Gap = -0.0040008 m VT = -0.142343427 Vs 12:13:40   Voltage Measurement, V = 0.000000 V 12:13:41 Temperature Measurement, Probe Number = 0, T = 20.582000 C 12:13:41 Temperature Measurement, Probe Number = 1, T = 21.530000 C 12:13:41 Magnet Test Parameter Measurements P0 = Main (A) , Value = 0.00 P1 = Walker_Meas_Junction, Value = 20.58 P2 = 3458_Meas_Junction, Value = 21.53 12:13:41 Integrated Voltage Measurement... 12:14:13 Voltage Offset Correction: v_off_ini = -0.000000987 V, v_off_fin = 0.000001271 V 12:14:13 Integrated Voltage Measurement Delta Gap = 0.0040012 m VT = 0.142388568 Vs 12:14:13 Integrated Voltage Measurement... 12:14:44 Voltage Offset Correction: v_off_ini = 0.000000725 V, v_off_fin = -0.000003086 V 12:14:44 Integrated Voltage Measurement Delta Gap = -0.0040006 m VT = -0.142332857 Vs 12:14:45   Voltage Measurement, V = 0.000000 V 12:14:45 Temperature Measurement, Probe Number = 0, T = 20.560000 C 12:14:46 Temperature Measurement, Probe Number = 1, T = 21.517000 C 12:14:46 Magnet Test Parameter Measurements P0 = Main (A) , Value = 0.00 P1 = Walker_Meas_Junction, Value = 20.56 P2 = 3458_Meas_Junction, Value = 21.52 12:14:46 Integrated Voltage Measurement... 12:15:16 Voltage Offset Correction: v_off_ini = -0.000001049 V, v_off_fin = 0.000001976 V 12:15:16 Integrated Voltage Measurement Delta Gap = 0.0040011 m VT = 0.142391072 Vs 12:15:16 Integrated Voltage Measurement... 12:15:49 Voltage Offset Correction: v_off_ini = 0.000000820 V, v_off_fin = -0.000003333 V 12:15:49 Integrated Voltage Measurement Delta Gap = -0.0040007 m VT = -0.142335441 Vs 12:15:50   Voltage Measurement, V = 0.000000 V 12:15:51 Temperature Measurement, Probe Number = 0, T = 20.555000 C 12:15:51 Temperature Measurement, Probe Number = 1, T = 21.520000 C 12:15:51 Magnet Test Parameter Measurements P0 = Main (A) , Value = 0.00 P1 = Walker_Meas_Junction, Value = 20.55 P2 = 3458_Meas_Junction, Value = 21.52 12:15:51 Integrated Voltage Measurement... 12:16:22 Voltage Offset Correction: v_off_ini = -0.000000802 V, v_off_fin = 0.000000905 V 12:16:22 Integrated Voltage Measurement Delta Gap = 0.0040012 m VT = 0.142386063 Vs 12:16:22 Integrated Voltage Measurement... 12:16:57 Voltage Offset Correction: v_off_ini = 0.000000850 V, v_off_fin = -0.000002256 V 12:16:57 Integrated Voltage Measurement Delta Gap = -0.0040009 m VT = -0.142343115 Vs 12:16:58   Voltage Measurement, V = 0.000000 V 12:16:59 Temperature Measurement, Probe Number = 0, T = 20.551000 C 12:17:00 Temperature Measurement, Probe Number = 1, T = 21.513000 C 12:17:00 Magnet Test Parameter Measurements P0 = Main (A) , Value = 0.00 P1 = Walker_Meas_Junction, Value = 20.55 P2 = 3458_Meas_Junction, Value = 21.51 12:17:00 Integrated Voltage Measurement... 12:17:31 Voltage Offset Correction: v_off_ini = -0.000000873 V, v_off_fin = 0.000001465 V 12:17:31 Integrated Voltage Measurement Delta Gap = 0.0040013 m VT = 0.142384382 Vs 12:17:31 Integrated Voltage Measurement... 12:18:02 Voltage Offset Correction: v_off_ini = 0.000000747 V, v_off_fin = -0.000003505 V 12:18:02 Integrated Voltage Measurement Delta Gap = -0.0040009 m VT = -0.142349964 Vs