SLAC Magnetic Measurements Date: 01-21-2021 Time: 10:43:02 Project: LCLS-II Magnet Type: Radiation_Detection_System Magnet Name: SXU_009_Cell_36_Upstairs Serial Number: PC33408 Run Number: 9 Measurement Device: Coil Operator: sda Comment: Coil Integral measurement upstairs, cable through switch Full Set 9 10:43:21 Ambient Measurement Device(s) Being Used, type = None 10:43:21 Magnet Voltage Measurement Device(s) Being Used, type = None 10:43:21 Temperature Measurement Device(s) Being Used, type = HPTHERM_HP34970 10:43:21 XMAG Device(s) Being Used, type = None 10:43:21 Ambient Measurement Device(s) Being Used, type = None 10:43:25   Voltage Measurement, V = 0.000000 V 10:43:26 Temperature Measurement, Probe Number = 0, T = 23.579000 C 10:43:27 Temperature Measurement, Probe Number = 1, T = 24.285000 C 10:43:27 Magnet Test Parameter Measurements P0 = Main (A) , Value = 0.00 P1 = Walker_Meas_Junction, Value = 23.58 P2 = 3458_Meas_Junction, Value = 24.29 10:43:27 Integrated Voltage Measurement... 10:43:53 Voltage Offset Correction: v_off_ini = 0.000003160 V, v_off_fin = 0.000004947 V 10:43:53 Integrated Voltage Measurement Delta Gap = 0.0039995 m VT = 0.143308550 Vs 10:43:53 Integrated Voltage Measurement... 10:44:28 Voltage Offset Correction: v_off_ini = 0.000003184 V, v_off_fin = -0.000000267 V 10:44:28 Integrated Voltage Measurement Delta Gap = -0.0040010 m VT = -0.143364221 Vs 10:44:29   Voltage Measurement, V = 0.000000 V 10:44:29 Temperature Measurement, Probe Number = 0, T = 23.576000 C 10:44:30 Temperature Measurement, Probe Number = 1, T = 24.276000 C 10:44:30 Magnet Test Parameter Measurements P0 = Main (A) , Value = 0.00 P1 = Walker_Meas_Junction, Value = 23.58 P2 = 3458_Meas_Junction, Value = 24.28 10:44:30 Integrated Voltage Measurement... 10:45:02 Voltage Offset Correction: v_off_ini = 0.000001082 V, v_off_fin = 0.000003955 V 10:45:02 Integrated Voltage Measurement Delta Gap = 0.0040013 m VT = 0.143395323 Vs 10:45:02 Integrated Voltage Measurement... 10:45:33 Voltage Offset Correction: v_off_ini = 0.000003394 V, v_off_fin = -0.000000575 V 10:45:33 Integrated Voltage Measurement Delta Gap = -0.0040009 m VT = -0.143352401 Vs 10:45:34   Voltage Measurement, V = 0.000000 V 10:45:34 Temperature Measurement, Probe Number = 0, T = 23.582000 C 10:45:35 Temperature Measurement, Probe Number = 1, T = 24.287000 C 10:45:35 Magnet Test Parameter Measurements P0 = Main (A) , Value = 0.00 P1 = Walker_Meas_Junction, Value = 23.58 P2 = 3458_Meas_Junction, Value = 24.29 10:45:35 Integrated Voltage Measurement... 10:46:06 Voltage Offset Correction: v_off_ini = 0.000001387 V, v_off_fin = 0.000003503 V 10:46:06 Integrated Voltage Measurement Delta Gap = 0.0040014 m VT = 0.143410230 Vs 10:46:06 Integrated Voltage Measurement... 10:46:37 Voltage Offset Correction: v_off_ini = 0.000003226 V, v_off_fin = -0.000001539 V 10:46:37 Integrated Voltage Measurement Delta Gap = -0.0040009 m VT = -0.143356662 Vs 10:46:38   Voltage Measurement, V = 0.000000 V 10:46:39 Temperature Measurement, Probe Number = 0, T = 23.588000 C 10:46:39 Temperature Measurement, Probe Number = 1, T = 24.271000 C 10:46:39 Magnet Test Parameter Measurements P0 = Main (A) , Value = 0.00 P1 = Walker_Meas_Junction, Value = 23.59 P2 = 3458_Meas_Junction, Value = 24.27 10:46:39 Integrated Voltage Measurement... 10:47:11 Voltage Offset Correction: v_off_ini = 0.000001456 V, v_off_fin = 0.000003870 V 10:47:11 Integrated Voltage Measurement Delta Gap = 0.0040012 m VT = 0.143391115 Vs 10:47:11 Integrated Voltage Measurement... 10:47:41 Voltage Offset Correction: v_off_ini = 0.000003513 V, v_off_fin = 0.000000000 V 10:47:41 Integrated Voltage Measurement Delta Gap = -0.0040009 m VT = -0.143353776 Vs 10:47:42   Voltage Measurement, V = 0.000000 V 10:47:43 Temperature Measurement, Probe Number = 0, T = 23.567000 C 10:47:44 Temperature Measurement, Probe Number = 1, T = 24.256000 C 10:47:44 Magnet Test Parameter Measurements P0 = Main (A) , Value = 0.00 P1 = Walker_Meas_Junction, Value = 23.57 P2 = 3458_Meas_Junction, Value = 24.26 10:47:44 Integrated Voltage Measurement... 10:48:14 Voltage Offset Correction: v_off_ini = 0.000001166 V, v_off_fin = 0.000004467 V 10:48:14 Integrated Voltage Measurement Delta Gap = 0.0040014 m VT = 0.143405825 Vs 10:48:14 Integrated Voltage Measurement... 10:48:45 Voltage Offset Correction: v_off_ini = 0.000003488 V, v_off_fin = -0.000000893 V 10:48:45 Integrated Voltage Measurement Delta Gap = -0.0040011 m VT = -0.143358323 Vs 10:48:46   Voltage Measurement, V = 0.000000 V 10:48:47 Temperature Measurement, Probe Number = 0, T = 23.557000 C 10:48:47 Temperature Measurement, Probe Number = 1, T = 24.273000 C 10:48:47 Magnet Test Parameter Measurements P0 = Main (A) , Value = 0.00 P1 = Walker_Meas_Junction, Value = 23.56 P2 = 3458_Meas_Junction, Value = 24.27 10:48:47 Integrated Voltage Measurement... 10:49:18 Voltage Offset Correction: v_off_ini = 0.000001253 V, v_off_fin = 0.000002931 V 10:49:18 Integrated Voltage Measurement Delta Gap = 0.0040016 m VT = 0.143417237 Vs 10:49:18 Integrated Voltage Measurement... 10:50:03 Voltage Offset Correction: v_off_ini = 0.000002964 V, v_off_fin = 0.000000396 V 10:50:03 Integrated Voltage Measurement Delta Gap = -0.0040013 m VT = -0.143362320 Vs 10:50:04   Voltage Measurement, V = 0.000000 V 10:50:05 Temperature Measurement, Probe Number = 0, T = 23.552000 C 10:50:05 Temperature Measurement, Probe Number = 1, T = 24.247000 C 10:50:05 Magnet Test Parameter Measurements P0 = Main (A) , Value = 0.00 P1 = Walker_Meas_Junction, Value = 23.55 P2 = 3458_Meas_Junction, Value = 24.25 10:50:05 Integrated Voltage Measurement... 10:50:36 Voltage Offset Correction: v_off_ini = 0.000001305 V, v_off_fin = 0.000003614 V 10:50:36 Integrated Voltage Measurement Delta Gap = 0.0040013 m VT = 0.143398675 Vs 10:50:36 Integrated Voltage Measurement... 10:51:13 Voltage Offset Correction: v_off_ini = 0.000003123 V, v_off_fin = -0.000002075 V 10:51:13 Integrated Voltage Measurement Delta Gap = -0.0040012 m VT = -0.143353258 Vs 10:51:14   Voltage Measurement, V = 0.000000 V 10:51:14 Temperature Measurement, Probe Number = 0, T = 23.526000 C 10:51:15 Temperature Measurement, Probe Number = 1, T = 24.209000 C 10:51:15 Magnet Test Parameter Measurements P0 = Main (A) , Value = 0.00 P1 = Walker_Meas_Junction, Value = 23.53 P2 = 3458_Meas_Junction, Value = 24.21 10:51:15 Integrated Voltage Measurement... 10:51:46 Voltage Offset Correction: v_off_ini = 0.000001106 V, v_off_fin = 0.000003612 V 10:51:46 Integrated Voltage Measurement Delta Gap = 0.0040016 m VT = 0.143414516 Vs 10:51:46 Integrated Voltage Measurement... 10:52:17 Voltage Offset Correction: v_off_ini = 0.000003466 V, v_off_fin = -0.000000448 V 10:52:17 Integrated Voltage Measurement Delta Gap = -0.0040011 m VT = -0.143352750 Vs 10:52:18   Voltage Measurement, V = 0.000000 V 10:52:18 Temperature Measurement, Probe Number = 0, T = 23.504000 C 10:52:19 Temperature Measurement, Probe Number = 1, T = 24.180000 C 10:52:19 Magnet Test Parameter Measurements P0 = Main (A) , Value = 0.00 P1 = Walker_Meas_Junction, Value = 23.50 P2 = 3458_Meas_Junction, Value = 24.18 10:52:19 Integrated Voltage Measurement... 10:52:50 Voltage Offset Correction: v_off_ini = 0.000001378 V, v_off_fin = 0.000004185 V 10:52:50 Integrated Voltage Measurement Delta Gap = 0.0040014 m VT = 0.143404803 Vs 10:52:50 Integrated Voltage Measurement... 10:53:22 Voltage Offset Correction: v_off_ini = 0.000003360 V, v_off_fin = -0.000000121 V 10:53:22 Integrated Voltage Measurement Delta Gap = -0.0040011 m VT = -0.143357183 Vs 10:53:23   Voltage Measurement, V = 0.000000 V 10:53:24 Temperature Measurement, Probe Number = 0, T = 23.499000 C 10:53:25 Temperature Measurement, Probe Number = 1, T = 24.171000 C 10:53:25 Magnet Test Parameter Measurements P0 = Main (A) , Value = 0.00 P1 = Walker_Meas_Junction, Value = 23.50 P2 = 3458_Meas_Junction, Value = 24.17 10:53:25 Integrated Voltage Measurement... 10:53:56 Voltage Offset Correction: v_off_ini = 0.000001404 V, v_off_fin = 0.000004106 V 10:53:56 Integrated Voltage Measurement Delta Gap = 0.0040014 m VT = 0.143400068 Vs 10:53:56 Integrated Voltage Measurement... 10:54:28 Voltage Offset Correction: v_off_ini = 0.000003489 V, v_off_fin = -0.000001214 V 10:54:28 Integrated Voltage Measurement Delta Gap = -0.0040011 m VT = -0.143348426 Vs