SLAC Magnetic Measurements Date: 10-20-2020 Time: 14:57:33 Project: LCLS-II Magnet Type: Radiation_Detection_System Magnet Name: SXU_006_Cell_26 Serial Number: PC33408 Run Number: 11 Measurement Device: Coil Operator: sda Comment: Coil Integral measurement in tunnel. Full Set 11 14:57:51 Ambient Measurement Device(s) Being Used, type = None 14:57:51 Magnet Voltage Measurement Device(s) Being Used, type = None 14:57:51 Temperature Measurement Device(s) Being Used, type = HPTHERM_HP34970 14:57:51 XMAG Device(s) Being Used, type = None 14:57:51 Ambient Measurement Device(s) Being Used, type = None 14:57:54   Voltage Measurement, V = 0.000000 V 14:57:55 Temperature Measurement, Probe Number = 0, T = 20.490000 C 14:57:56 Temperature Measurement, Probe Number = 1, T = 21.750000 C 14:57:56 Magnet Test Parameter Measurements P0 = Main (A) , Value = 0.00 P1 = Walker_Meas_Junction, Value = 20.49 P2 = 3458_Meas_Junction, Value = 21.75 14:57:56 Integrated Voltage Measurement... 14:58:23 Voltage Offset Correction: v_off_ini = -0.000000458 V, v_off_fin = 0.000000550 V 14:58:23 Integrated Voltage Measurement Delta Gap = 0.0040015 m VT = 0.144505698 Vs 14:58:23 Integrated Voltage Measurement... 14:58:55 Voltage Offset Correction: v_off_ini = 0.000000066 V, v_off_fin = -0.000001468 V 14:58:55 Integrated Voltage Measurement Delta Gap = -0.0040005 m VT = -0.144419416 Vs 14:58:56   Voltage Measurement, V = 0.000000 V 14:58:57 Temperature Measurement, Probe Number = 0, T = 20.476000 C 14:58:57 Temperature Measurement, Probe Number = 1, T = 21.734000 C 14:58:57 Magnet Test Parameter Measurements P0 = Main (A) , Value = 0.00 P1 = Walker_Meas_Junction, Value = 20.48 P2 = 3458_Meas_Junction, Value = 21.73 14:58:57 Integrated Voltage Measurement... 14:59:30 Voltage Offset Correction: v_off_ini = -0.000000674 V, v_off_fin = 0.000001659 V 14:59:30 Integrated Voltage Measurement Delta Gap = 0.0040005 m VT = 0.144455365 Vs 14:59:30 Integrated Voltage Measurement... 15:00:02 Voltage Offset Correction: v_off_ini = -0.000000225 V, v_off_fin = -0.000002109 V 15:00:02 Integrated Voltage Measurement Delta Gap = -0.0040003 m VT = -0.144417485 Vs 15:00:03   Voltage Measurement, V = 0.000000 V 15:00:04 Temperature Measurement, Probe Number = 0, T = 20.508000 C 15:00:05 Temperature Measurement, Probe Number = 1, T = 21.730000 C 15:00:05 Magnet Test Parameter Measurements P0 = Main (A) , Value = 0.00 P1 = Walker_Meas_Junction, Value = 20.51 P2 = 3458_Meas_Junction, Value = 21.73 15:00:05 Integrated Voltage Measurement... 15:00:40 Voltage Offset Correction: v_off_ini = -0.000001088 V, v_off_fin = 0.000000181 V 15:00:40 Integrated Voltage Measurement Delta Gap = 0.0040005 m VT = 0.144447001 Vs 15:00:40 Integrated Voltage Measurement... 15:01:14 Voltage Offset Correction: v_off_ini = -0.000000004 V, v_off_fin = -0.000001186 V 15:01:14 Integrated Voltage Measurement Delta Gap = -0.0040004 m VT = -0.144431910 Vs 15:01:15   Voltage Measurement, V = 0.000000 V 15:01:16 Temperature Measurement, Probe Number = 0, T = 20.528000 C 15:01:17 Temperature Measurement, Probe Number = 1, T = 21.715000 C 15:01:17 Magnet Test Parameter Measurements P0 = Main (A) , Value = 0.00 P1 = Walker_Meas_Junction, Value = 20.53 P2 = 3458_Meas_Junction, Value = 21.71 15:01:17 Integrated Voltage Measurement... 15:01:51 Voltage Offset Correction: v_off_ini = -0.000000977 V, v_off_fin = 0.000000037 V 15:01:51 Integrated Voltage Measurement Delta Gap = 0.0040006 m VT = 0.144457165 Vs 15:01:51 Integrated Voltage Measurement... 15:02:24 Voltage Offset Correction: v_off_ini = -0.000000113 V, v_off_fin = -0.000002406 V 15:02:24 Integrated Voltage Measurement Delta Gap = -0.0040005 m VT = -0.144431875 Vs 15:02:25   Voltage Measurement, V = 0.000000 V 15:02:26 Temperature Measurement, Probe Number = 0, T = 20.512000 C 15:02:27 Temperature Measurement, Probe Number = 1, T = 21.724000 C 15:02:27 Magnet Test Parameter Measurements P0 = Main (A) , Value = 0.00 P1 = Walker_Meas_Junction, Value = 20.51 P2 = 3458_Meas_Junction, Value = 21.72 15:02:27 Integrated Voltage Measurement... 15:02:59 Voltage Offset Correction: v_off_ini = -0.000001226 V, v_off_fin = -0.000001481 V 15:02:59 Integrated Voltage Measurement Delta Gap = 0.0040005 m VT = 0.144457103 Vs 15:02:59 Integrated Voltage Measurement... 15:03:37 Voltage Offset Correction: v_off_ini = -0.000000038 V, v_off_fin = -0.000001229 V 15:03:37 Integrated Voltage Measurement Delta Gap = -0.0040005 m VT = -0.144431131 Vs 15:03:38   Voltage Measurement, V = 0.000000 V 15:03:39 Temperature Measurement, Probe Number = 0, T = 20.497000 C 15:03:40 Temperature Measurement, Probe Number = 1, T = 21.738000 C 15:03:40 Magnet Test Parameter Measurements P0 = Main (A) , Value = 0.00 P1 = Walker_Meas_Junction, Value = 20.50 P2 = 3458_Meas_Junction, Value = 21.74 15:03:40 Integrated Voltage Measurement... 15:04:11 Voltage Offset Correction: v_off_ini = -0.000001119 V, v_off_fin = -0.000000503 V 15:04:11 Integrated Voltage Measurement Delta Gap = 0.0040004 m VT = 0.144454742 Vs 15:04:11 Integrated Voltage Measurement... 15:04:43 Voltage Offset Correction: v_off_ini = -0.000000073 V, v_off_fin = -0.000002265 V 15:04:43 Integrated Voltage Measurement Delta Gap = -0.0040004 m VT = -0.144431481 Vs 15:04:44   Voltage Measurement, V = 0.000000 V 15:04:45 Temperature Measurement, Probe Number = 0, T = 20.483000 C 15:04:46 Temperature Measurement, Probe Number = 1, T = 21.743000 C 15:04:46 Magnet Test Parameter Measurements P0 = Main (A) , Value = 0.00 P1 = Walker_Meas_Junction, Value = 20.48 P2 = 3458_Meas_Junction, Value = 21.74 15:04:46 Integrated Voltage Measurement... 15:05:22 Voltage Offset Correction: v_off_ini = -0.000001169 V, v_off_fin = -0.000000239 V 15:05:22 Integrated Voltage Measurement Delta Gap = 0.0040006 m VT = 0.144455021 Vs 15:05:22 Integrated Voltage Measurement... 15:06:03 Voltage Offset Correction: v_off_ini = -0.000000148 V, v_off_fin = -0.000002516 V 15:06:03 Integrated Voltage Measurement Delta Gap = -0.0040005 m VT = -0.144428626 Vs 15:06:04   Voltage Measurement, V = 0.000000 V 15:06:05 Temperature Measurement, Probe Number = 0, T = 20.463000 C 15:06:06 Temperature Measurement, Probe Number = 1, T = 21.743000 C 15:06:06 Magnet Test Parameter Measurements P0 = Main (A) , Value = 0.00 P1 = Walker_Meas_Junction, Value = 20.46 P2 = 3458_Meas_Junction, Value = 21.74 15:06:06 Integrated Voltage Measurement... 15:06:39 Voltage Offset Correction: v_off_ini = -0.000001108 V, v_off_fin = 0.000000820 V 15:06:39 Integrated Voltage Measurement Delta Gap = 0.0040006 m VT = 0.144463177 Vs 15:06:39 Integrated Voltage Measurement... 15:07:12 Voltage Offset Correction: v_off_ini = -0.000000203 V, v_off_fin = -0.000002449 V 15:07:12 Integrated Voltage Measurement Delta Gap = -0.0040004 m VT = -0.144438732 Vs 15:07:13   Voltage Measurement, V = 0.000000 V 15:07:13 Temperature Measurement, Probe Number = 0, T = 20.457000 C 15:07:14 Temperature Measurement, Probe Number = 1, T = 21.732000 C 15:07:14 Magnet Test Parameter Measurements P0 = Main (A) , Value = 0.00 P1 = Walker_Meas_Junction, Value = 20.46 P2 = 3458_Meas_Junction, Value = 21.73 15:07:14 Integrated Voltage Measurement... 15:07:48 Voltage Offset Correction: v_off_ini = -0.000001126 V, v_off_fin = -0.000000425 V 15:07:48 Integrated Voltage Measurement Delta Gap = 0.0040005 m VT = 0.144452105 Vs 15:07:48 Integrated Voltage Measurement... 15:08:19 Voltage Offset Correction: v_off_ini = -0.000000079 V, v_off_fin = -0.000002756 V 15:08:19 Integrated Voltage Measurement Delta Gap = -0.0040005 m VT = -0.144431022 Vs 15:08:20   Voltage Measurement, V = 0.000000 V 15:08:21 Temperature Measurement, Probe Number = 0, T = 20.460000 C 15:08:21 Temperature Measurement, Probe Number = 1, T = 21.736000 C 15:08:21 Magnet Test Parameter Measurements P0 = Main (A) , Value = 0.00 P1 = Walker_Meas_Junction, Value = 20.46 P2 = 3458_Meas_Junction, Value = 21.74 15:08:21 Integrated Voltage Measurement... 15:08:54 Voltage Offset Correction: v_off_ini = -0.000001252 V, v_off_fin = 0.000000271 V 15:08:54 Integrated Voltage Measurement Delta Gap = 0.0040005 m VT = 0.144449969 Vs 15:08:54 Integrated Voltage Measurement... 15:09:25 Voltage Offset Correction: v_off_ini = 0.000000042 V, v_off_fin = -0.000002775 V 15:09:25 Integrated Voltage Measurement Delta Gap = -0.0040005 m VT = -0.144432469 Vs