SLAC Magnetic Measurements Date: 02-09-2018 Time: 13:02:42 Project: LCLS2 Magnet Type: Undulator Magnet Name: SXU_Day6 Serial Number: 000 Run Number: 60 Measurement Device: PXI4464 Operator: jeb Comment: fs=100 Hz tb=0.2 s ta=6.0 s Coil attached with tape. Offset corr: TRUE USING THE FILTER SR650 fc=0.3 Hz, LP 40dB/dec, G=1, Dynamic Reserve Mode Undulator Gap Integral vs Gap Change Distance the gap was moved = 0.004601 m Time Gap Start Gap Change Voltage Integral (m) (m) (Vs) ---------- ----------- ------------ ------------------ 13:03:11 0.007299 0.0046012 -0.1923915 13:03:34 0.011900 -0.0046004 0.1921005 13:04:01 0.007300 0.0046005 -0.1924721 13:04:23 0.011900 -0.0046005 0.1921063 13:04:50 0.007299 0.0046007 -0.1925296 13:05:12 0.011900 -0.0046005 0.1920712 13:05:40 0.007299 0.0046007 -0.1924825 13:06:02 0.011900 -0.0046005 0.1921023 13:06:29 0.007300 0.0046006 -0.1924368 13:06:51 0.011900 -0.0046005 0.1921121 13:07:18 0.007299 0.0046006 -0.1924470 13:07:40 0.011900 -0.0046005 0.1920592 13:08:07 0.007300 0.0046006 -0.1924179 13:08:29 0.011900 -0.0046005 0.1920890 13:08:56 0.007299 0.0046007 -0.1924624 13:09:18 0.011900 -0.0046005 0.1920587 13:09:45 0.007299 0.0046007 -0.1924380 13:10:08 0.011900 -0.0046006 0.1920988 13:10:35 0.007299 0.0046007 -0.1924534 13:10:56 0.011900 -0.0046005 0.1921112 ------------------------- Positive movement: VT = -0.1924531 +- 0.0000376 (Vs) (RSD = -0.0001953) Negative movement: VT = 0.1920909 +- 0.0000206 (Vs) (RSD = 0.0001073) Difference between averages: VT_pos+VT_neg = -0.0003622 (Vs) (-0.1884 %)