SLAC Magnetic Measurements Date: 10-04-2022 Time: 10:40:26 Project: LCLS-II-HE Device Type: Undulator Data Type: Serial Number: HE_SXU_000 DATASET Number: 1 Tuning Stage: Tuning Run Number: 75 Measurement Type: Long Coil Scans Operator: YL Comment: Skew Quad, 4H Center position: X0=-0.002000, Y0=0.000000 Nominal Gap=15.00 Scan Raw value Integral Nominal Actual Pre-measurement Gap Actual Post-measurement Gap Number Volt*sec T*m | T*m^2 Gap US - mm US Raw DS - mm DS Raw US - mm US Raw DS - mm DS Raw --------------------------------------------------------------------------------------------------------------------------------------- 0 1.817E-06 25.947E-06 15.00000 15.00140 0 15.00015 0 15.00140 0 15.00020 0 1 1.549E-06 22.126E-06 15.00000 15.00145 0 15.00015 0 15.00145 0 15.00015 0 2 1.648E-06 23.535E-06 15.00000 15.00140 0 15.00015 0 15.00140 0 15.00015 0 3 1.553E-06 22.180E-06 15.00000 15.00145 0 15.00015 0 15.00140 0 15.00020 0 4 1.540E-06 22.001E-06 15.00000 15.00145 0 15.00020 0 15.00145 0 15.00015 0