SLAC Magnetic Measurements Date: 07-15-2026 Time: 12:08:27 Project: LCLS-II-HE Magnet Type: Phase Shifter Magnet Name: HE_SXPS_16335 Dataset Number: 1 Run Purpose: Tuning Run Type: Run Number: 1 Measurement Device: Stretched Wire Operator: YL Comment: 12:09:26 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 12:09:28 Motion Parameters, X Controller, Axis 1 acc = 0.001000 m/s/s dcl = 0.001000 m/s/s vel = 0.001000 m/s pos = 0.000000 m kp = 188.600000 ki = 820.000000 kd = 491.400000 db = 0.000000 12:09:30 Motion Parameters, X Controller, Axis 2 acc = 0.001000 m/s/s dcl = 0.001000 m/s/s vel = 0.001000 m/s pos = 0.000000 m kp = 188.600000 ki = 820.000000 kd = 491.400000 db = 0.000000 12:09:31 Motion Parameters, Y Controller, Axis 1 acc = 0.001000 m/s/s dcl = 0.001000 m/s/s vel = 0.001000 m/s pos = 0.000000 m kp = 188.600000 ki = 820.000000 kd = 491.400000 db = 0.000000 12:09:33 Motion Parameters, Y Controller, Axis 2 acc = 0.001000 m/s/s dcl = 0.001000 m/s/s vel = 0.001000 m/s pos = 0.000000 m kp = 188.600000 ki = 820.000000 kd = 491.400000 db = 0.000000 12:11:09 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 12:11:13 Integrated Voltage Measurement... 12:11:22 Voltage Offset Correction: v_off_ini = 1977.505 uV, v_off_fin = 1955.709 uV 12:11:22 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -215.499 uVs 12:11:22 Integrated Voltage Measurement... 12:11:31 Voltage Offset Correction: v_off_ini = 1974.811 uV, v_off_fin = 1963.840 uV 12:11:31 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 174.566 uVs 12:11:31 Integrated Voltage Measurement... 12:11:40 Voltage Offset Correction: v_off_ini = 1972.234 uV, v_off_fin = 1980.027 uV 12:11:40 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -217.668 uVs 12:11:40 Integrated Voltage Measurement... 12:11:49 Voltage Offset Correction: v_off_ini = 1968.889 uV, v_off_fin = 1983.222 uV 12:11:49 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 199.409 uVs 12:11:49 Integrated Voltage Measurement... 12:11:58 Voltage Offset Correction: v_off_ini = 1994.971 uV, v_off_fin = 1981.652 uV 12:11:58 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -218.417 uVs 12:11:58 Integrated Voltage Measurement... 12:12:07 Voltage Offset Correction: v_off_ini = 1967.940 uV, v_off_fin = 1974.931 uV 12:12:07 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 196.781 uVs 12:12:12 I1X Average and RMS of Measurements VT = -203.723 +- 6.157 uVs I1X = 40.746 +- 1.231 uTm 12:12:14 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 12:12:18 Integrated Voltage Measurement... 12:12:27 Voltage Offset Correction: v_off_ini = 1984.403 uV, v_off_fin = 1978.979 uV 12:12:27 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -242.914 uVs 12:12:27 Integrated Voltage Measurement... 12:12:36 Voltage Offset Correction: v_off_ini = 1978.924 uV, v_off_fin = 2014.537 uV 12:12:36 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 260.381 uVs 12:12:36 Integrated Voltage Measurement... 12:12:45 Voltage Offset Correction: v_off_ini = 1983.373 uV, v_off_fin = 1973.803 uV 12:12:45 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -246.481 uVs 12:12:45 Integrated Voltage Measurement... 12:12:54 Voltage Offset Correction: v_off_ini = 1990.786 uV, v_off_fin = 1995.947 uV 12:12:54 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 271.036 uVs 12:12:54 Integrated Voltage Measurement... 12:13:03 Voltage Offset Correction: v_off_ini = 2003.234 uV, v_off_fin = 1996.449 uV 12:13:03 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -254.153 uVs 12:13:03 Integrated Voltage Measurement... 12:13:12 Voltage Offset Correction: v_off_ini = 2003.941 uV, v_off_fin = 2014.503 uV 12:13:12 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 261.449 uVs 12:13:16 I1Y Average and RMS of Measurements VT = -256.069 +- 3.151 uVs I1Y = -51.214 +- 0.630 uTm 12:13:21 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 12:13:24 Integrated Voltage Measurement... 12:13:33 Voltage Offset Correction: v_off_ini = 1975.388 uV, v_off_fin = 1981.260 uV 12:13:33 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -156.681 uVs 12:13:33 Integrated Voltage Measurement... 12:13:42 Voltage Offset Correction: v_off_ini = 1984.505 uV, v_off_fin = 2003.551 uV 12:13:42 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 155.363 uVs 12:13:42 Integrated Voltage Measurement... 12:13:51 Voltage Offset Correction: v_off_ini = 1999.429 uV, v_off_fin = 1995.606 uV 12:13:51 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -191.901 uVs 12:13:51 Integrated Voltage Measurement... 12:14:00 Voltage Offset Correction: v_off_ini = 2027.657 uV, v_off_fin = 2029.953 uV 12:14:00 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 149.009 uVs 12:14:00 Integrated Voltage Measurement... 12:14:09 Voltage Offset Correction: v_off_ini = 1995.038 uV, v_off_fin = 1989.453 uV 12:14:09 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -129.449 uVs 12:14:09 Integrated Voltage Measurement... 12:14:18 Voltage Offset Correction: v_off_ini = 1985.154 uV, v_off_fin = 1994.583 uV 12:14:18 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 179.938 uVs 12:14:20 I2X Average and RMS of Measurements VT = -160.390 +- 7.137 uVs I2X = 14.756 +- 0.657 uTm^2 12:14:26 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 12:14:28 Integrated Voltage Measurement... 12:14:37 Voltage Offset Correction: v_off_ini = 1974.180 uV, v_off_fin = 1989.713 uV 12:14:37 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -331.751 uVs 12:14:37 Integrated Voltage Measurement... 12:14:46 Voltage Offset Correction: v_off_ini = 1995.321 uV, v_off_fin = 2004.706 uV 12:14:46 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 333.184 uVs 12:14:46 Integrated Voltage Measurement... 12:14:55 Voltage Offset Correction: v_off_ini = 1990.634 uV, v_off_fin = 1999.521 uV 12:14:55 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -300.837 uVs 12:14:55 Integrated Voltage Measurement... 12:15:04 Voltage Offset Correction: v_off_ini = 1978.062 uV, v_off_fin = 1998.914 uV 12:15:04 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 320.269 uVs 12:15:04 Integrated Voltage Measurement... 12:15:13 Voltage Offset Correction: v_off_ini = 2002.353 uV, v_off_fin = 2000.180 uV 12:15:13 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -300.994 uVs 12:15:13 Integrated Voltage Measurement... 12:15:22 Voltage Offset Correction: v_off_ini = 1997.386 uV, v_off_fin = 2004.493 uV 12:15:22 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 319.750 uVs 12:15:25 I2Y Average and RMS of Measurements VT = -317.797 +- 10.374 uVs I2Y = -29.237 +- 0.954 uTm^2 12:15:35 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 12:15:40 Integrated Voltage Measurement... 12:15:49 Voltage Offset Correction: v_off_ini = 2037.347 uV, v_off_fin = 2002.795 uV 12:15:49 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -211.484 uVs 12:15:49 Integrated Voltage Measurement... 12:15:58 Voltage Offset Correction: v_off_ini = 1990.895 uV, v_off_fin = 2006.153 uV 12:15:58 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 166.862 uVs 12:15:58 Integrated Voltage Measurement... 12:16:07 Voltage Offset Correction: v_off_ini = 2011.377 uV, v_off_fin = 2000.522 uV 12:16:07 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -211.035 uVs 12:16:07 Integrated Voltage Measurement... 12:16:16 Voltage Offset Correction: v_off_ini = 2012.433 uV, v_off_fin = 2003.902 uV 12:16:16 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 189.979 uVs 12:16:16 Integrated Voltage Measurement... 12:16:25 Voltage Offset Correction: v_off_ini = 1988.949 uV, v_off_fin = 1972.371 uV 12:16:25 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -196.238 uVs 12:16:25 Integrated Voltage Measurement... 12:16:34 Voltage Offset Correction: v_off_ini = 1993.034 uV, v_off_fin = 1997.283 uV 12:16:34 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 169.398 uVs 12:16:38 I1X Average and RMS of Measurements VT = -190.833 +- 7.316 uVs I1X = 38.167 +- 1.463 uTm 12:16:40 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 12:16:45 Integrated Voltage Measurement... 12:16:54 Voltage Offset Correction: v_off_ini = 1976.360 uV, v_off_fin = 1973.604 uV 12:16:54 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = -237.271 uVs 12:16:54 Integrated Voltage Measurement... 12:17:03 Voltage Offset Correction: v_off_ini = 1998.783 uV, v_off_fin = 2001.783 uV 12:17:03 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 262.105 uVs 12:17:03 Integrated Voltage Measurement... 12:17:12 Voltage Offset Correction: v_off_ini = 1989.959 uV, v_off_fin = 1996.640 uV 12:17:12 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -254.033 uVs 12:17:12 Integrated Voltage Measurement... 12:17:21 Voltage Offset Correction: v_off_ini = 2023.275 uV, v_off_fin = 2022.298 uV 12:17:21 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 253.396 uVs 12:17:21 Integrated Voltage Measurement... 12:17:30 Voltage Offset Correction: v_off_ini = 2026.399 uV, v_off_fin = 1996.659 uV 12:17:30 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -244.409 uVs 12:17:30 Integrated Voltage Measurement... 12:17:39 Voltage Offset Correction: v_off_ini = 2010.264 uV, v_off_fin = 2040.672 uV 12:17:39 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 276.192 uVs 12:17:43 I1Y Average and RMS of Measurements VT = -254.568 +- 4.374 uVs I1Y = -50.914 +- 0.874 uTm 12:17:48 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 12:17:50 Integrated Voltage Measurement... 12:17:59 Voltage Offset Correction: v_off_ini = 2015.665 uV, v_off_fin = 1998.379 uV 12:17:59 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -139.026 uVs 12:17:59 Integrated Voltage Measurement... 12:18:08 Voltage Offset Correction: v_off_ini = 1991.763 uV, v_off_fin = 2007.021 uV 12:18:08 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 166.323 uVs 12:18:08 Integrated Voltage Measurement... 12:18:17 Voltage Offset Correction: v_off_ini = 1975.483 uV, v_off_fin = 1980.616 uV 12:18:17 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -117.399 uVs 12:18:17 Integrated Voltage Measurement... 12:18:26 Voltage Offset Correction: v_off_ini = 1979.793 uV, v_off_fin = 2003.363 uV 12:18:26 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 136.438 uVs 12:18:26 Integrated Voltage Measurement... 12:18:35 Voltage Offset Correction: v_off_ini = 1985.439 uV, v_off_fin = 1982.267 uV 12:18:35 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -126.564 uVs 12:18:36 Integrated Voltage Measurement... 12:18:44 Voltage Offset Correction: v_off_ini = 1983.936 uV, v_off_fin = 1984.053 uV 12:18:45 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 143.703 uVs 12:18:47 I2X Average and RMS of Measurements VT = -138.242 +- 10.742 uVs I2X = 12.718 +- 0.988 uTm^2 12:18:52 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 12:18:55 Integrated Voltage Measurement... 12:19:04 Voltage Offset Correction: v_off_ini = 1984.248 uV, v_off_fin = 2001.686 uV 12:19:04 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -284.131 uVs 12:19:04 Integrated Voltage Measurement... 12:19:13 Voltage Offset Correction: v_off_ini = 2008.503 uV, v_off_fin = 2001.505 uV 12:19:13 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 294.146 uVs 12:19:13 Integrated Voltage Measurement... 12:19:22 Voltage Offset Correction: v_off_ini = 1967.688 uV, v_off_fin = 1973.218 uV 12:19:22 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -280.818 uVs 12:19:22 Integrated Voltage Measurement... 12:19:31 Voltage Offset Correction: v_off_ini = 1991.156 uV, v_off_fin = 1996.763 uV 12:19:31 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 291.565 uVs 12:19:31 Integrated Voltage Measurement... 12:19:40 Voltage Offset Correction: v_off_ini = 1966.883 uV, v_off_fin = 1986.960 uV 12:19:40 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -281.509 uVs 12:19:40 Integrated Voltage Measurement... 12:19:49 Voltage Offset Correction: v_off_ini = 1997.251 uV, v_off_fin = 1987.182 uV 12:19:49 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 287.809 uVs 12:19:51 I2Y Average and RMS of Measurements VT = -286.663 +- 1.859 uVs I2Y = -26.373 +- 0.171 uTm^2 12:20:02 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 12:20:06 Integrated Voltage Measurement... 12:20:15 Voltage Offset Correction: v_off_ini = 1997.799 uV, v_off_fin = 1974.711 uV 12:20:15 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.4998 mm dXY = -0.4999 mm VT = -151.674 uVs 12:20:15 Integrated Voltage Measurement... 12:20:24 Voltage Offset Correction: v_off_ini = 1983.455 uV, v_off_fin = 2006.530 uV 12:20:24 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.4998 mm dXY = 0.4999 mm VT = 152.012 uVs 12:20:24 Integrated Voltage Measurement... 12:20:33 Voltage Offset Correction: v_off_ini = 2008.654 uV, v_off_fin = 1998.261 uV 12:20:33 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -170.908 uVs 12:20:33 Integrated Voltage Measurement... 12:20:42 Voltage Offset Correction: v_off_ini = 1991.709 uV, v_off_fin = 1995.758 uV 12:20:42 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 176.125 uVs 12:20:42 Integrated Voltage Measurement... 12:20:51 Voltage Offset Correction: v_off_ini = 1989.111 uV, v_off_fin = 2004.570 uV 12:20:51 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -174.762 uVs 12:20:51 Integrated Voltage Measurement... 12:21:00 Voltage Offset Correction: v_off_ini = 1969.079 uV, v_off_fin = 2002.804 uV 12:21:00 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 188.358 uVs 12:21:04 I1X Average and RMS of Measurements VT = -168.973 +- 12.550 uVs I1X = 33.798 +- 2.506 uTm 12:21:07 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 12:21:11 Integrated Voltage Measurement... 12:21:20 Voltage Offset Correction: v_off_ini = 2027.651 uV, v_off_fin = 1996.799 uV 12:21:20 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = -253.469 uVs 12:21:20 Integrated Voltage Measurement... 12:21:29 Voltage Offset Correction: v_off_ini = 2011.378 uV, v_off_fin = 2008.307 uV 12:21:29 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 256.060 uVs 12:21:29 Integrated Voltage Measurement... 12:21:38 Voltage Offset Correction: v_off_ini = 1996.833 uV, v_off_fin = 1981.000 uV 12:21:38 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -242.092 uVs 12:21:38 Integrated Voltage Measurement... 12:21:47 Voltage Offset Correction: v_off_ini = 2004.963 uV, v_off_fin = 2031.602 uV 12:21:47 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 224.213 uVs 12:21:47 Integrated Voltage Measurement... 12:21:56 Voltage Offset Correction: v_off_ini = 1975.411 uV, v_off_fin = 1945.511 uV 12:21:56 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -231.218 uVs 12:21:56 Integrated Voltage Measurement... 12:22:05 Voltage Offset Correction: v_off_ini = 1977.618 uV, v_off_fin = 2004.860 uV 12:22:05 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 265.220 uVs 12:22:09 I1Y Average and RMS of Measurements VT = -245.379 +- 9.049 uVs I1Y = -49.076 +- 1.810 uTm 12:22:14 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 12:22:17 Integrated Voltage Measurement... 12:22:26 Voltage Offset Correction: v_off_ini = 1966.128 uV, v_off_fin = 1975.259 uV 12:22:26 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -134.179 uVs 12:22:26 Integrated Voltage Measurement... 12:22:35 Voltage Offset Correction: v_off_ini = 1978.742 uV, v_off_fin = 1997.437 uV 12:22:35 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 140.648 uVs 12:22:35 Integrated Voltage Measurement... 12:22:44 Voltage Offset Correction: v_off_ini = 1990.615 uV, v_off_fin = 1973.165 uV 12:22:44 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -129.740 uVs 12:22:44 Integrated Voltage Measurement... 12:22:53 Voltage Offset Correction: v_off_ini = 1977.288 uV, v_off_fin = 1973.958 uV 12:22:53 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 100.036 uVs 12:22:53 Integrated Voltage Measurement... 12:23:02 Voltage Offset Correction: v_off_ini = 1970.579 uV, v_off_fin = 1973.576 uV 12:23:02 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -132.759 uVs 12:23:02 Integrated Voltage Measurement... 12:23:11 Voltage Offset Correction: v_off_ini = 1979.939 uV, v_off_fin = 1951.736 uV 12:23:11 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 131.999 uVs 12:23:13 I2X Average and RMS of Measurements VT = -128.227 +- 9.654 uVs I2X = 11.797 +- 0.888 uTm^2 12:23:18 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 12:23:21 Integrated Voltage Measurement... 12:23:30 Voltage Offset Correction: v_off_ini = 1988.210 uV, v_off_fin = 2019.161 uV 12:23:30 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -282.538 uVs 12:23:30 Integrated Voltage Measurement... 12:23:39 Voltage Offset Correction: v_off_ini = 2031.420 uV, v_off_fin = 2013.508 uV 12:23:39 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 277.808 uVs 12:23:39 Integrated Voltage Measurement... 12:23:48 Voltage Offset Correction: v_off_ini = 2001.828 uV, v_off_fin = 2010.431 uV 12:23:48 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -255.253 uVs 12:23:48 Integrated Voltage Measurement... 12:23:57 Voltage Offset Correction: v_off_ini = 2023.658 uV, v_off_fin = 2014.574 uV 12:23:57 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 278.068 uVs 12:23:57 Integrated Voltage Measurement... 12:24:06 Voltage Offset Correction: v_off_ini = 1994.456 uV, v_off_fin = 1995.252 uV 12:24:06 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -267.591 uVs 12:24:06 Integrated Voltage Measurement... 12:24:15 Voltage Offset Correction: v_off_ini = 2013.310 uV, v_off_fin = 2029.206 uV 12:24:15 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 270.117 uVs 12:24:17 I2Y Average and RMS of Measurements VT = -271.896 +- 5.921 uVs I2Y = -25.014 +- 0.545 uTm^2 12:24:28 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 12:24:32 Integrated Voltage Measurement... 12:24:41 Voltage Offset Correction: v_off_ini = 2050.601 uV, v_off_fin = 2031.950 uV 12:24:41 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -146.070 uVs 12:24:41 Integrated Voltage Measurement... 12:24:50 Voltage Offset Correction: v_off_ini = 2015.756 uV, v_off_fin = 2012.222 uV 12:24:50 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 165.835 uVs 12:24:50 Integrated Voltage Measurement... 12:24:59 Voltage Offset Correction: v_off_ini = 2036.852 uV, v_off_fin = 2009.629 uV 12:24:59 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -179.434 uVs 12:24:59 Integrated Voltage Measurement... 12:25:08 Voltage Offset Correction: v_off_ini = 2004.266 uV, v_off_fin = 2027.293 uV 12:25:08 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 174.748 uVs 12:25:08 Integrated Voltage Measurement... 12:25:17 Voltage Offset Correction: v_off_ini = 2040.421 uV, v_off_fin = 2040.505 uV 12:25:17 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -147.107 uVs 12:25:17 Integrated Voltage Measurement... 12:25:26 Voltage Offset Correction: v_off_ini = 2061.401 uV, v_off_fin = 2069.801 uV 12:25:26 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 153.185 uVs 12:25:31 I1X Average and RMS of Measurements VT = -161.063 +- 11.579 uVs I1X = 32.213 +- 2.317 uTm 12:25:33 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 12:25:37 Integrated Voltage Measurement... 12:25:46 Voltage Offset Correction: v_off_ini = 2068.710 uV, v_off_fin = 2064.586 uV 12:25:46 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -228.800 uVs 12:25:46 Integrated Voltage Measurement... 12:25:55 Voltage Offset Correction: v_off_ini = 2071.082 uV, v_off_fin = 2090.811 uV 12:25:55 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 221.506 uVs 12:25:55 Integrated Voltage Measurement... 12:26:04 Voltage Offset Correction: v_off_ini = 2066.522 uV, v_off_fin = 2074.309 uV 12:26:04 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -209.746 uVs 12:26:04 Integrated Voltage Measurement... 12:26:13 Voltage Offset Correction: v_off_ini = 2073.904 uV, v_off_fin = 2079.045 uV 12:26:13 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 226.660 uVs 12:26:13 Integrated Voltage Measurement... 12:26:22 Voltage Offset Correction: v_off_ini = 2064.142 uV, v_off_fin = 2056.630 uV 12:26:22 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -236.071 uVs 12:26:22 Integrated Voltage Measurement... 12:26:31 Voltage Offset Correction: v_off_ini = 2064.772 uV, v_off_fin = 2078.854 uV 12:26:31 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 242.818 uVs 12:26:35 I1Y Average and RMS of Measurements VT = -227.600 +- 8.843 uVs I1Y = -45.516 +- 1.768 uTm 12:26:41 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 12:26:43 Integrated Voltage Measurement... 12:26:52 Voltage Offset Correction: v_off_ini = 2088.702 uV, v_off_fin = 2111.027 uV 12:26:52 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -80.641 uVs 12:26:52 Integrated Voltage Measurement... 12:27:01 Voltage Offset Correction: v_off_ini = 2133.718 uV, v_off_fin = 2114.308 uV 12:27:01 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 89.432 uVs 12:27:01 Integrated Voltage Measurement... 12:27:10 Voltage Offset Correction: v_off_ini = 2105.334 uV, v_off_fin = 2083.072 uV 12:27:10 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -85.746 uVs 12:27:10 Integrated Voltage Measurement... 12:27:19 Voltage Offset Correction: v_off_ini = 2091.262 uV, v_off_fin = 2103.556 uV 12:27:19 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 126.766 uVs 12:27:19 Integrated Voltage Measurement... 12:27:28 Voltage Offset Correction: v_off_ini = 2102.117 uV, v_off_fin = 2107.488 uV 12:27:28 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -116.003 uVs 12:27:28 Integrated Voltage Measurement... 12:27:37 Voltage Offset Correction: v_off_ini = 2115.996 uV, v_off_fin = 2137.107 uV 12:27:37 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 65.883 uVs 12:27:40 I2X Average and RMS of Measurements VT = -94.079 +- 8.942 uVs I2X = 8.655 +- 0.823 uTm^2 12:27:45 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 12:27:47 Integrated Voltage Measurement... 12:27:56 Voltage Offset Correction: v_off_ini = 2113.671 uV, v_off_fin = 2116.187 uV 12:27:56 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -320.067 uVs 12:27:56 Integrated Voltage Measurement... 12:28:05 Voltage Offset Correction: v_off_ini = 2129.045 uV, v_off_fin = 2139.608 uV 12:28:05 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 267.504 uVs 12:28:06 Integrated Voltage Measurement... 12:28:14 Voltage Offset Correction: v_off_ini = 2113.213 uV, v_off_fin = 2171.758 uV 12:28:15 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -255.013 uVs 12:28:15 Integrated Voltage Measurement... 12:28:24 Voltage Offset Correction: v_off_ini = 2145.500 uV, v_off_fin = 2131.980 uV 12:28:24 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 268.579 uVs 12:28:24 Integrated Voltage Measurement... 12:28:33 Voltage Offset Correction: v_off_ini = 2123.660 uV, v_off_fin = 2141.910 uV 12:28:33 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = -266.840 uVs 12:28:33 Integrated Voltage Measurement... 12:28:42 Voltage Offset Correction: v_off_ini = 2136.472 uV, v_off_fin = 2110.613 uV 12:28:42 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 260.628 uVs 12:28:44 I2Y Average and RMS of Measurements VT = -273.105 +- 14.645 uVs I2Y = -25.126 +- 1.347 uTm^2 12:28:55 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 12:28:59 Integrated Voltage Measurement... 12:29:08 Voltage Offset Correction: v_off_ini = 2137.412 uV, v_off_fin = 2140.771 uV 12:29:08 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -107.227 uVs 12:29:08 Integrated Voltage Measurement... 12:29:17 Voltage Offset Correction: v_off_ini = 2106.816 uV, v_off_fin = 2104.246 uV 12:29:17 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 135.552 uVs 12:29:17 Integrated Voltage Measurement... 12:29:26 Voltage Offset Correction: v_off_ini = 2120.967 uV, v_off_fin = 2112.680 uV 12:29:26 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -152.037 uVs 12:29:26 Integrated Voltage Measurement... 12:29:35 Voltage Offset Correction: v_off_ini = 2145.401 uV, v_off_fin = 2143.723 uV 12:29:35 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 140.292 uVs 12:29:35 Integrated Voltage Measurement... 12:29:44 Voltage Offset Correction: v_off_ini = 2127.765 uV, v_off_fin = 2135.070 uV 12:29:44 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -139.992 uVs 12:29:44 Integrated Voltage Measurement... 12:29:53 Voltage Offset Correction: v_off_ini = 2154.780 uV, v_off_fin = 2165.734 uV 12:29:53 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 140.121 uVs 12:29:58 I1X Average and RMS of Measurements VT = -135.870 +- 10.539 uVs I1X = 27.174 +- 2.108 uTm 12:30:00 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 12:30:04 Integrated Voltage Measurement... 12:30:13 Voltage Offset Correction: v_off_ini = 2126.431 uV, v_off_fin = 2133.008 uV 12:30:13 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -223.294 uVs 12:30:13 Integrated Voltage Measurement... 12:30:22 Voltage Offset Correction: v_off_ini = 2141.875 uV, v_off_fin = 2141.404 uV 12:30:22 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 224.189 uVs 12:30:22 Integrated Voltage Measurement... 12:30:31 Voltage Offset Correction: v_off_ini = 2139.852 uV, v_off_fin = 2140.556 uV 12:30:31 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -239.078 uVs 12:30:31 Integrated Voltage Measurement... 12:30:40 Voltage Offset Correction: v_off_ini = 2170.435 uV, v_off_fin = 2182.571 uV 12:30:40 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 216.727 uVs 12:30:40 Integrated Voltage Measurement... 12:30:49 Voltage Offset Correction: v_off_ini = 2166.536 uV, v_off_fin = 2148.264 uV 12:30:49 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -217.436 uVs 12:30:49 Integrated Voltage Measurement... 12:30:58 Voltage Offset Correction: v_off_ini = 2126.962 uV, v_off_fin = 2143.297 uV 12:30:58 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 237.900 uVs 12:31:03 I1Y Average and RMS of Measurements VT = -226.437 +- 1.909 uVs I1Y = -45.286 +- 0.382 uTm 12:31:08 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 12:31:10 Integrated Voltage Measurement... 12:31:19 Voltage Offset Correction: v_off_ini = 2137.312 uV, v_off_fin = 2146.662 uV 12:31:19 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -73.918 uVs 12:31:19 Integrated Voltage Measurement... 12:31:28 Voltage Offset Correction: v_off_ini = 2150.932 uV, v_off_fin = 2152.304 uV 12:31:28 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 82.564 uVs 12:31:28 Integrated Voltage Measurement... 12:31:37 Voltage Offset Correction: v_off_ini = 2142.745 uV, v_off_fin = 2147.335 uV 12:31:37 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -68.849 uVs 12:31:37 Integrated Voltage Measurement... 12:31:46 Voltage Offset Correction: v_off_ini = 2143.268 uV, v_off_fin = 2156.903 uV 12:31:46 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 92.335 uVs 12:31:46 Integrated Voltage Measurement... 12:31:55 Voltage Offset Correction: v_off_ini = 2152.454 uV, v_off_fin = 2151.883 uV 12:31:55 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -96.814 uVs 12:31:55 Integrated Voltage Measurement... 12:32:04 Voltage Offset Correction: v_off_ini = 2164.912 uV, v_off_fin = 2163.149 uV 12:32:04 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 79.998 uVs 12:32:07 I2X Average and RMS of Measurements VT = -82.413 +- 4.345 uVs I2X = 7.582 +- 0.400 uTm^2 12:32:12 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 12:32:14 Integrated Voltage Measurement... 12:32:23 Voltage Offset Correction: v_off_ini = 2158.284 uV, v_off_fin = 2132.879 uV 12:32:23 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -256.698 uVs 12:32:23 Integrated Voltage Measurement... 12:32:32 Voltage Offset Correction: v_off_ini = 2154.923 uV, v_off_fin = 2147.386 uV 12:32:32 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 252.151 uVs 12:32:32 Integrated Voltage Measurement... 12:32:41 Voltage Offset Correction: v_off_ini = 2132.871 uV, v_off_fin = 2145.717 uV 12:32:41 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -267.989 uVs 12:32:41 Integrated Voltage Measurement... 12:32:50 Voltage Offset Correction: v_off_ini = 2136.279 uV, v_off_fin = 2150.686 uV 12:32:50 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 245.030 uVs 12:32:51 Integrated Voltage Measurement... 12:32:59 Voltage Offset Correction: v_off_ini = 2148.637 uV, v_off_fin = 2154.760 uV 12:33:00 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -246.804 uVs 12:33:00 Integrated Voltage Measurement... 12:33:08 Voltage Offset Correction: v_off_ini = 2159.730 uV, v_off_fin = 2159.807 uV 12:33:09 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 256.570 uVs 12:33:11 I2Y Average and RMS of Measurements VT = -254.207 +- 1.975 uVs I2Y = -23.387 +- 0.182 uTm^2 12:33:22 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 12:33:26 Integrated Voltage Measurement... 12:33:35 Voltage Offset Correction: v_off_ini = 2189.800 uV, v_off_fin = 2194.136 uV 12:33:35 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -120.526 uVs 12:33:35 Integrated Voltage Measurement... 12:33:44 Voltage Offset Correction: v_off_ini = 2224.466 uV, v_off_fin = 2207.265 uV 12:33:44 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 99.687 uVs 12:33:44 Integrated Voltage Measurement... 12:33:53 Voltage Offset Correction: v_off_ini = 2176.583 uV, v_off_fin = 2179.579 uV 12:33:53 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -124.160 uVs 12:33:53 Integrated Voltage Measurement... 12:34:02 Voltage Offset Correction: v_off_ini = 2196.442 uV, v_off_fin = 2196.200 uV 12:34:02 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 92.620 uVs 12:34:02 Integrated Voltage Measurement... 12:34:11 Voltage Offset Correction: v_off_ini = 2201.940 uV, v_off_fin = 2200.594 uV 12:34:11 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -128.378 uVs 12:34:11 Integrated Voltage Measurement... 12:34:20 Voltage Offset Correction: v_off_ini = 2187.427 uV, v_off_fin = 2188.974 uV 12:34:20 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 115.531 uVs 12:34:24 I1X Average and RMS of Measurements VT = -113.484 +- 6.031 uVs I1X = 22.698 +- 1.207 uTm 12:34:27 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 12:34:31 Integrated Voltage Measurement... 12:34:40 Voltage Offset Correction: v_off_ini = 2212.297 uV, v_off_fin = 2214.612 uV 12:34:40 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = -199.213 uVs 12:34:40 Integrated Voltage Measurement... 12:34:49 Voltage Offset Correction: v_off_ini = 2205.337 uV, v_off_fin = 2234.628 uV 12:34:49 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 197.878 uVs 12:34:49 Integrated Voltage Measurement... 12:34:58 Voltage Offset Correction: v_off_ini = 2227.901 uV, v_off_fin = 2210.163 uV 12:34:58 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -194.841 uVs 12:34:58 Integrated Voltage Measurement... 12:35:07 Voltage Offset Correction: v_off_ini = 2220.168 uV, v_off_fin = 2255.736 uV 12:35:07 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 223.659 uVs 12:35:07 Integrated Voltage Measurement... 12:35:16 Voltage Offset Correction: v_off_ini = 2244.258 uV, v_off_fin = 2226.505 uV 12:35:16 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -242.534 uVs 12:35:16 Integrated Voltage Measurement... 12:35:25 Voltage Offset Correction: v_off_ini = 2220.151 uV, v_off_fin = 2250.718 uV 12:35:25 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 191.212 uVs 12:35:29 I1Y Average and RMS of Measurements VT = -208.223 +- 7.517 uVs I1Y = -41.644 +- 1.502 uTm 12:35:34 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 12:35:37 Integrated Voltage Measurement... 12:35:46 Voltage Offset Correction: v_off_ini = 2245.675 uV, v_off_fin = 2249.098 uV 12:35:46 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -35.159 uVs 12:35:46 Integrated Voltage Measurement... 12:35:55 Voltage Offset Correction: v_off_ini = 2245.626 uV, v_off_fin = 2263.828 uV 12:35:55 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 41.267 uVs 12:35:55 Integrated Voltage Measurement... 12:36:04 Voltage Offset Correction: v_off_ini = 2252.848 uV, v_off_fin = 2246.780 uV 12:36:04 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -58.370 uVs 12:36:04 Integrated Voltage Measurement... 12:36:13 Voltage Offset Correction: v_off_ini = 2234.189 uV, v_off_fin = 2256.030 uV 12:36:13 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 74.056 uVs 12:36:13 Integrated Voltage Measurement... 12:36:22 Voltage Offset Correction: v_off_ini = 2255.597 uV, v_off_fin = 2243.061 uV 12:36:22 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -78.049 uVs 12:36:22 Integrated Voltage Measurement... 12:36:31 Voltage Offset Correction: v_off_ini = 2257.155 uV, v_off_fin = 2268.216 uV 12:36:31 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 46.954 uVs 12:36:33 I2X Average and RMS of Measurements VT = -55.642 +- 12.417 uVs I2X = 5.119 +- 1.143 uTm^2 12:36:38 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 12:36:41 Integrated Voltage Measurement... 12:36:50 Voltage Offset Correction: v_off_ini = 2269.369 uV, v_off_fin = 2271.987 uV 12:36:50 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -247.159 uVs 12:36:50 Integrated Voltage Measurement... 12:36:59 Voltage Offset Correction: v_off_ini = 2265.121 uV, v_off_fin = 2280.623 uV 12:36:59 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 249.508 uVs 12:36:59 Integrated Voltage Measurement... 12:37:08 Voltage Offset Correction: v_off_ini = 2288.683 uV, v_off_fin = 2291.616 uV 12:37:08 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -227.694 uVs 12:37:08 Integrated Voltage Measurement... 12:37:17 Voltage Offset Correction: v_off_ini = 2310.931 uV, v_off_fin = 2324.636 uV 12:37:17 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 246.070 uVs 12:37:17 Integrated Voltage Measurement... 12:37:26 Voltage Offset Correction: v_off_ini = 2311.216 uV, v_off_fin = 2300.836 uV 12:37:26 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -229.591 uVs 12:37:26 Integrated Voltage Measurement... 12:37:35 Voltage Offset Correction: v_off_ini = 2318.128 uV, v_off_fin = 2331.540 uV 12:37:35 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 239.135 uVs 12:37:38 I2Y Average and RMS of Measurements VT = -239.860 +- 6.080 uVs I2Y = -22.067 +- 0.559 uTm^2 12:37:48 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 12:37:53 Integrated Voltage Measurement... 12:38:02 Voltage Offset Correction: v_off_ini = 2293.144 uV, v_off_fin = 2273.594 uV 12:38:02 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -104.687 uVs 12:38:02 Integrated Voltage Measurement... 12:38:11 Voltage Offset Correction: v_off_ini = 2276.261 uV, v_off_fin = 2284.953 uV 12:38:11 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 124.487 uVs 12:38:11 Integrated Voltage Measurement... 12:38:20 Voltage Offset Correction: v_off_ini = 2295.414 uV, v_off_fin = 2301.728 uV 12:38:20 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -95.247 uVs 12:38:20 Integrated Voltage Measurement... 12:38:29 Voltage Offset Correction: v_off_ini = 2287.346 uV, v_off_fin = 2290.902 uV 12:38:29 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 121.851 uVs 12:38:29 Integrated Voltage Measurement... 12:38:38 Voltage Offset Correction: v_off_ini = 2301.930 uV, v_off_fin = 2300.108 uV 12:38:38 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -83.447 uVs 12:38:38 Integrated Voltage Measurement... 12:38:47 Voltage Offset Correction: v_off_ini = 2290.293 uV, v_off_fin = 2294.538 uV 12:38:47 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = 101.223 uVs 12:38:51 I1X Average and RMS of Measurements VT = -105.157 +- 9.396 uVs I1X = 21.033 +- 1.879 uTm 12:38:53 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 12:38:58 Integrated Voltage Measurement... 12:39:07 Voltage Offset Correction: v_off_ini = 2317.514 uV, v_off_fin = 2310.085 uV 12:39:07 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -205.015 uVs 12:39:07 Integrated Voltage Measurement... 12:39:16 Voltage Offset Correction: v_off_ini = 2302.405 uV, v_off_fin = 2318.756 uV 12:39:16 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 215.922 uVs 12:39:16 Integrated Voltage Measurement... 12:39:25 Voltage Offset Correction: v_off_ini = 2321.928 uV, v_off_fin = 2278.560 uV 12:39:25 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -231.202 uVs 12:39:25 Integrated Voltage Measurement... 12:39:34 Voltage Offset Correction: v_off_ini = 2302.046 uV, v_off_fin = 2301.660 uV 12:39:34 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 214.960 uVs 12:39:34 Integrated Voltage Measurement... 12:39:43 Voltage Offset Correction: v_off_ini = 2299.333 uV, v_off_fin = 2298.523 uV 12:39:43 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -193.310 uVs 12:39:43 Integrated Voltage Measurement... 12:39:52 Voltage Offset Correction: v_off_ini = 2334.875 uV, v_off_fin = 2319.512 uV 12:39:52 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 201.289 uVs 12:39:56 I1Y Average and RMS of Measurements VT = -210.283 +- 10.526 uVs I1Y = -42.057 +- 2.105 uTm 12:40:01 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 12:40:03 Integrated Voltage Measurement... 12:40:12 Voltage Offset Correction: v_off_ini = 2308.071 uV, v_off_fin = 2317.048 uV 12:40:12 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -38.537 uVs 12:40:13 Integrated Voltage Measurement... 12:40:21 Voltage Offset Correction: v_off_ini = 2310.542 uV, v_off_fin = 2317.919 uV 12:40:22 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 32.459 uVs 12:40:22 Integrated Voltage Measurement... 12:40:31 Voltage Offset Correction: v_off_ini = 2339.567 uV, v_off_fin = 2315.077 uV 12:40:31 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -69.241 uVs 12:40:31 Integrated Voltage Measurement... 12:40:40 Voltage Offset Correction: v_off_ini = 2309.209 uV, v_off_fin = 2328.382 uV 12:40:40 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 62.836 uVs 12:40:40 Integrated Voltage Measurement... 12:40:49 Voltage Offset Correction: v_off_ini = 2333.427 uV, v_off_fin = 2310.174 uV 12:40:49 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -38.355 uVs 12:40:49 Integrated Voltage Measurement... 12:40:58 Voltage Offset Correction: v_off_ini = 2329.025 uV, v_off_fin = 2343.467 uV 12:40:58 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 31.081 uVs 12:41:00 I2X Average and RMS of Measurements VT = -45.418 +- 14.584 uVs I2X = 4.179 +- 1.342 uTm^2 12:41:05 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 12:41:08 Integrated Voltage Measurement... 12:41:17 Voltage Offset Correction: v_off_ini = 2358.471 uV, v_off_fin = 2323.998 uV 12:41:17 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -238.585 uVs 12:41:17 Integrated Voltage Measurement... 12:41:26 Voltage Offset Correction: v_off_ini = 2329.109 uV, v_off_fin = 2326.289 uV 12:41:26 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 200.062 uVs 12:41:26 Integrated Voltage Measurement... 12:41:35 Voltage Offset Correction: v_off_ini = 2327.380 uV, v_off_fin = 2344.645 uV 12:41:35 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -244.215 uVs 12:41:35 Integrated Voltage Measurement... 12:41:44 Voltage Offset Correction: v_off_ini = 2330.010 uV, v_off_fin = 2338.028 uV 12:41:44 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 216.231 uVs 12:41:44 Integrated Voltage Measurement... 12:41:53 Voltage Offset Correction: v_off_ini = 2360.001 uV, v_off_fin = 2349.630 uV 12:41:53 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -225.437 uVs 12:41:53 Integrated Voltage Measurement... 12:42:02 Voltage Offset Correction: v_off_ini = 2350.913 uV, v_off_fin = 2352.627 uV 12:42:02 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 235.731 uVs 12:42:04 I2Y Average and RMS of Measurements VT = -226.710 +- 5.225 uVs I2Y = -20.857 +- 0.481 uTm^2 12:42:15 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 12:42:19 Integrated Voltage Measurement... 12:42:28 Voltage Offset Correction: v_off_ini = 2359.335 uV, v_off_fin = 2313.622 uV 12:42:28 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -96.895 uVs 12:42:28 Integrated Voltage Measurement... 12:42:37 Voltage Offset Correction: v_off_ini = 2317.240 uV, v_off_fin = 2336.294 uV 12:42:37 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 103.350 uVs 12:42:37 Integrated Voltage Measurement... 12:42:46 Voltage Offset Correction: v_off_ini = 2343.870 uV, v_off_fin = 2341.812 uV 12:42:46 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -86.312 uVs 12:42:46 Integrated Voltage Measurement... 12:42:55 Voltage Offset Correction: v_off_ini = 2359.638 uV, v_off_fin = 2348.407 uV 12:42:55 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 106.718 uVs 12:42:55 Integrated Voltage Measurement... 12:43:04 Voltage Offset Correction: v_off_ini = 2346.605 uV, v_off_fin = 2348.720 uV 12:43:04 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -104.398 uVs 12:43:04 Integrated Voltage Measurement... 12:43:13 Voltage Offset Correction: v_off_ini = 2348.297 uV, v_off_fin = 2366.345 uV 12:43:13 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 75.363 uVs 12:43:18 I1X Average and RMS of Measurements VT = -95.506 +- 4.242 uVs I1X = 19.101 +- 0.848 uTm 12:43:20 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 12:43:24 Integrated Voltage Measurement... 12:43:33 Voltage Offset Correction: v_off_ini = 2378.144 uV, v_off_fin = 2371.005 uV 12:43:33 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = -229.584 uVs 12:43:33 Integrated Voltage Measurement... 12:43:42 Voltage Offset Correction: v_off_ini = 2383.954 uV, v_off_fin = 2391.300 uV 12:43:42 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 239.033 uVs 12:43:42 Integrated Voltage Measurement... 12:43:51 Voltage Offset Correction: v_off_ini = 2401.434 uV, v_off_fin = 2376.913 uV 12:43:51 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -211.856 uVs 12:43:51 Integrated Voltage Measurement... 12:44:00 Voltage Offset Correction: v_off_ini = 2389.847 uV, v_off_fin = 2389.915 uV 12:44:00 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 188.948 uVs 12:44:00 Integrated Voltage Measurement... 12:44:09 Voltage Offset Correction: v_off_ini = 2381.416 uV, v_off_fin = 2375.069 uV 12:44:09 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -192.917 uVs 12:44:09 Integrated Voltage Measurement... 12:44:18 Voltage Offset Correction: v_off_ini = 2388.251 uV, v_off_fin = 2423.082 uV 12:44:18 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 219.497 uVs 12:44:22 I1Y Average and RMS of Measurements VT = -213.639 +- 14.806 uVs I1Y = -42.729 +- 2.962 uTm 12:44:28 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 12:44:30 Integrated Voltage Measurement... 12:44:39 Voltage Offset Correction: v_off_ini = 2388.718 uV, v_off_fin = 2394.115 uV 12:44:39 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -23.340 uVs 12:44:39 Integrated Voltage Measurement... 12:44:48 Voltage Offset Correction: v_off_ini = 2408.571 uV, v_off_fin = 2426.019 uV 12:44:48 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 26.027 uVs 12:44:48 Integrated Voltage Measurement... 12:44:57 Voltage Offset Correction: v_off_ini = 2400.981 uV, v_off_fin = 2385.355 uV 12:44:57 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -27.396 uVs 12:44:57 Integrated Voltage Measurement... 12:45:06 Voltage Offset Correction: v_off_ini = 2402.762 uV, v_off_fin = 2436.120 uV 12:45:06 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 34.005 uVs 12:45:06 Integrated Voltage Measurement... 12:45:15 Voltage Offset Correction: v_off_ini = 2434.008 uV, v_off_fin = 2406.156 uV 12:45:15 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -31.317 uVs 12:45:15 Integrated Voltage Measurement... 12:45:24 Voltage Offset Correction: v_off_ini = 2409.628 uV, v_off_fin = 2417.869 uV 12:45:24 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 9.697 uVs 12:45:27 I2X Average and RMS of Measurements VT = -25.297 +- 4.184 uVs I2X = 2.327 +- 0.385 uTm^2 12:45:32 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 12:45:34 Integrated Voltage Measurement... 12:45:43 Voltage Offset Correction: v_off_ini = 2405.268 uV, v_off_fin = 2411.634 uV 12:45:43 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -213.967 uVs 12:45:43 Integrated Voltage Measurement... 12:45:52 Voltage Offset Correction: v_off_ini = 2413.206 uV, v_off_fin = 2421.874 uV 12:45:52 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 227.303 uVs 12:45:52 Integrated Voltage Measurement... 12:46:01 Voltage Offset Correction: v_off_ini = 2409.095 uV, v_off_fin = 2418.687 uV 12:46:01 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -226.727 uVs 12:46:01 Integrated Voltage Measurement... 12:46:10 Voltage Offset Correction: v_off_ini = 2422.743 uV, v_off_fin = 2427.605 uV 12:46:10 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 236.041 uVs 12:46:10 Integrated Voltage Measurement... 12:46:19 Voltage Offset Correction: v_off_ini = 2448.949 uV, v_off_fin = 2436.813 uV 12:46:19 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -257.061 uVs 12:46:19 Integrated Voltage Measurement... 12:46:28 Voltage Offset Correction: v_off_ini = 2413.661 uV, v_off_fin = 2422.275 uV 12:46:28 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 220.933 uVs 12:46:31 I2Y Average and RMS of Measurements VT = -230.339 +- 7.533 uVs I2Y = -21.191 +- 0.693 uTm^2 12:46:42 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 12:46:46 Integrated Voltage Measurement... 12:46:55 Voltage Offset Correction: v_off_ini = 2445.164 uV, v_off_fin = 2422.371 uV 12:46:55 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -49.279 uVs 12:46:55 Integrated Voltage Measurement... 12:47:04 Voltage Offset Correction: v_off_ini = 2442.826 uV, v_off_fin = 2447.342 uV 12:47:04 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 51.828 uVs 12:47:04 Integrated Voltage Measurement... 12:47:13 Voltage Offset Correction: v_off_ini = 2405.217 uV, v_off_fin = 2419.252 uV 12:47:13 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -83.364 uVs 12:47:13 Integrated Voltage Measurement... 12:47:22 Voltage Offset Correction: v_off_ini = 2413.169 uV, v_off_fin = 2412.639 uV 12:47:22 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 81.911 uVs 12:47:22 Integrated Voltage Measurement... 12:47:31 Voltage Offset Correction: v_off_ini = 2417.740 uV, v_off_fin = 2423.810 uV 12:47:31 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -66.459 uVs 12:47:31 Integrated Voltage Measurement... 12:47:40 Voltage Offset Correction: v_off_ini = 2427.832 uV, v_off_fin = 2433.942 uV 12:47:40 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 62.840 uVs 12:47:44 I1X Average and RMS of Measurements VT = -65.947 +- 13.130 uVs I1X = 13.189 +- 2.626 uTm 12:47:47 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 12:47:51 Integrated Voltage Measurement... 12:48:00 Voltage Offset Correction: v_off_ini = 2417.943 uV, v_off_fin = 2400.562 uV 12:48:00 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = -201.494 uVs 12:48:00 Integrated Voltage Measurement... 12:48:09 Voltage Offset Correction: v_off_ini = 2426.331 uV, v_off_fin = 2426.552 uV 12:48:09 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 205.438 uVs 12:48:09 Integrated Voltage Measurement... 12:48:18 Voltage Offset Correction: v_off_ini = 2410.091 uV, v_off_fin = 2391.815 uV 12:48:18 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -222.333 uVs 12:48:18 Integrated Voltage Measurement... 12:48:27 Voltage Offset Correction: v_off_ini = 2415.170 uV, v_off_fin = 2413.246 uV 12:48:27 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 197.143 uVs 12:48:27 Integrated Voltage Measurement... 12:48:36 Voltage Offset Correction: v_off_ini = 2425.396 uV, v_off_fin = 2404.504 uV 12:48:36 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -202.080 uVs 12:48:36 Integrated Voltage Measurement... 12:48:45 Voltage Offset Correction: v_off_ini = 2429.147 uV, v_off_fin = 2424.137 uV 12:48:45 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 201.950 uVs 12:48:49 I1Y Average and RMS of Measurements VT = -205.073 +- 3.351 uVs I1Y = -41.015 +- 0.670 uTm 12:48:54 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 12:48:57 Integrated Voltage Measurement... 12:49:06 Voltage Offset Correction: v_off_ini = 2408.490 uV, v_off_fin = 2386.539 uV 12:49:06 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 0.256 uVs 12:49:06 Integrated Voltage Measurement... 12:49:15 Voltage Offset Correction: v_off_ini = 2393.812 uV, v_off_fin = 2402.284 uV 12:49:15 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 25.890 uVs 12:49:15 Integrated Voltage Measurement... 12:49:24 Voltage Offset Correction: v_off_ini = 2399.850 uV, v_off_fin = 2370.295 uV 12:49:24 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 7.006 uVs 12:49:24 Integrated Voltage Measurement... 12:49:33 Voltage Offset Correction: v_off_ini = 2400.365 uV, v_off_fin = 2416.123 uV 12:49:33 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 33.156 uVs 12:49:33 Integrated Voltage Measurement... 12:49:42 Voltage Offset Correction: v_off_ini = 2418.335 uV, v_off_fin = 2402.000 uV 12:49:42 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -6.955 uVs 12:49:42 Integrated Voltage Measurement... 12:49:51 Voltage Offset Correction: v_off_ini = 2409.233 uV, v_off_fin = 2416.459 uV 12:49:51 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 18.739 uVs 12:49:53 I2X Average and RMS of Measurements VT = -12.913 +- 0.115 uVs I2X = 1.188 +- 0.011 uTm^2 12:49:58 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 12:50:01 Integrated Voltage Measurement... 12:50:10 Voltage Offset Correction: v_off_ini = 2402.473 uV, v_off_fin = 2377.469 uV 12:50:10 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -250.331 uVs 12:50:10 Integrated Voltage Measurement... 12:50:19 Voltage Offset Correction: v_off_ini = 2380.735 uV, v_off_fin = 2379.866 uV 12:50:19 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 229.413 uVs 12:50:19 Integrated Voltage Measurement... 12:50:28 Voltage Offset Correction: v_off_ini = 2375.686 uV, v_off_fin = 2363.304 uV 12:50:28 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -222.484 uVs 12:50:28 Integrated Voltage Measurement... 12:50:37 Voltage Offset Correction: v_off_ini = 2357.674 uV, v_off_fin = 2393.351 uV 12:50:37 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 223.797 uVs 12:50:37 Integrated Voltage Measurement... 12:50:46 Voltage Offset Correction: v_off_ini = 2387.432 uV, v_off_fin = 2386.176 uV 12:50:46 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -228.777 uVs 12:50:46 Integrated Voltage Measurement... 12:50:55 Voltage Offset Correction: v_off_ini = 2383.853 uV, v_off_fin = 2377.538 uV 12:50:55 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 215.736 uVs 12:50:57 I2Y Average and RMS of Measurements VT = -228.423 +- 8.104 uVs I2Y = -21.015 +- 0.746 uTm^2 12:51:08 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 12:51:12 Integrated Voltage Measurement... 12:51:21 Voltage Offset Correction: v_off_ini = 2414.058 uV, v_off_fin = 2394.044 uV 12:51:21 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -65.701 uVs 12:51:21 Integrated Voltage Measurement... 12:51:30 Voltage Offset Correction: v_off_ini = 2402.969 uV, v_off_fin = 2405.927 uV 12:51:30 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = 59.151 uVs 12:51:30 Integrated Voltage Measurement... 12:51:39 Voltage Offset Correction: v_off_ini = 2395.925 uV, v_off_fin = 2392.967 uV 12:51:39 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -58.910 uVs 12:51:39 Integrated Voltage Measurement... 12:51:48 Voltage Offset Correction: v_off_ini = 2391.241 uV, v_off_fin = 2392.866 uV 12:51:48 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 61.689 uVs 12:51:48 Integrated Voltage Measurement... 12:51:57 Voltage Offset Correction: v_off_ini = 2391.449 uV, v_off_fin = 2390.284 uV 12:51:57 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -60.616 uVs 12:51:57 Integrated Voltage Measurement... 12:52:06 Voltage Offset Correction: v_off_ini = 2390.706 uV, v_off_fin = 2393.342 uV 12:52:06 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 55.665 uVs 12:52:11 I1X Average and RMS of Measurements VT = -60.289 +- 1.749 uVs I1X = 12.059 +- 0.350 uTm 12:52:13 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 12:52:17 Integrated Voltage Measurement... 12:52:26 Voltage Offset Correction: v_off_ini = 2416.434 uV, v_off_fin = 2415.432 uV 12:52:26 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = -201.327 uVs 12:52:26 Integrated Voltage Measurement... 12:52:35 Voltage Offset Correction: v_off_ini = 2393.162 uV, v_off_fin = 2417.903 uV 12:52:35 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 204.818 uVs 12:52:35 Integrated Voltage Measurement... 12:52:44 Voltage Offset Correction: v_off_ini = 2404.139 uV, v_off_fin = 2418.203 uV 12:52:44 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -219.289 uVs 12:52:44 Integrated Voltage Measurement... 12:52:53 Voltage Offset Correction: v_off_ini = 2439.680 uV, v_off_fin = 2436.472 uV 12:52:53 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 185.533 uVs 12:52:53 Integrated Voltage Measurement... 12:53:02 Voltage Offset Correction: v_off_ini = 2434.370 uV, v_off_fin = 2406.137 uV 12:53:02 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -194.251 uVs 12:53:02 Integrated Voltage Measurement... 12:53:11 Voltage Offset Correction: v_off_ini = 2425.905 uV, v_off_fin = 2417.073 uV 12:53:11 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 176.025 uVs 12:53:15 I1Y Average and RMS of Measurements VT = -196.874 +- 8.303 uVs I1Y = -39.374 +- 1.661 uTm 12:53:21 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 12:53:23 Integrated Voltage Measurement... 12:53:32 Voltage Offset Correction: v_off_ini = 2385.145 uV, v_off_fin = 2407.112 uV 12:53:32 Integrated Voltage Measurement dXYus = -1.0001 mm, dXYds = 0.0000 mm dXY = -1.0001 mm VT = 19.308 uVs 12:53:32 Integrated Voltage Measurement... 12:53:41 Voltage Offset Correction: v_off_ini = 2416.368 uV, v_off_fin = 2421.690 uV 12:53:41 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -17.179 uVs 12:53:41 Integrated Voltage Measurement... 12:53:50 Voltage Offset Correction: v_off_ini = 2420.759 uV, v_off_fin = 2414.136 uV 12:53:50 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -1.217 uVs 12:53:50 Integrated Voltage Measurement... 12:53:59 Voltage Offset Correction: v_off_ini = 2433.712 uV, v_off_fin = 2438.778 uV 12:53:59 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -26.609 uVs 12:53:59 Integrated Voltage Measurement... 12:54:08 Voltage Offset Correction: v_off_ini = 2439.494 uV, v_off_fin = 2421.137 uV 12:54:08 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -3.562 uVs 12:54:08 Integrated Voltage Measurement... 12:54:17 Voltage Offset Correction: v_off_ini = 2445.931 uV, v_off_fin = 2441.881 uV 12:54:17 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -12.691 uVs 12:54:20 I2X Average and RMS of Measurements VT = 11.835 +- 5.617 uVs I2X = -1.089 +- 0.517 uTm^2 12:54:25 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 12:54:27 Integrated Voltage Measurement... 12:54:36 Voltage Offset Correction: v_off_ini = 2410.666 uV, v_off_fin = 2428.260 uV 12:54:36 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -213.490 uVs 12:54:36 Integrated Voltage Measurement... 12:54:45 Voltage Offset Correction: v_off_ini = 2434.990 uV, v_off_fin = 2430.451 uV 12:54:45 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 211.548 uVs 12:54:45 Integrated Voltage Measurement... 12:54:54 Voltage Offset Correction: v_off_ini = 2428.794 uV, v_off_fin = 2407.781 uV 12:54:54 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -217.784 uVs 12:54:54 Integrated Voltage Measurement... 12:55:03 Voltage Offset Correction: v_off_ini = 2401.063 uV, v_off_fin = 2419.849 uV 12:55:03 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 216.161 uVs 12:55:03 Integrated Voltage Measurement... 12:55:12 Voltage Offset Correction: v_off_ini = 2428.389 uV, v_off_fin = 2438.434 uV 12:55:12 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -225.455 uVs 12:55:13 Integrated Voltage Measurement... 12:55:21 Voltage Offset Correction: v_off_ini = 2428.724 uV, v_off_fin = 2425.486 uV 12:55:22 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 194.922 uVs 12:55:24 I2Y Average and RMS of Measurements VT = -213.226 +- 2.815 uVs I2Y = -19.617 +- 0.259 uTm^2 12:55:35 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 12:55:39 Integrated Voltage Measurement... 12:55:48 Voltage Offset Correction: v_off_ini = 2415.548 uV, v_off_fin = 2409.168 uV 12:55:48 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -38.474 uVs 12:55:48 Integrated Voltage Measurement... 12:55:57 Voltage Offset Correction: v_off_ini = 2417.760 uV, v_off_fin = 2443.554 uV 12:55:57 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = 49.726 uVs 12:55:57 Integrated Voltage Measurement... 12:56:06 Voltage Offset Correction: v_off_ini = 2407.511 uV, v_off_fin = 2450.600 uV 12:56:06 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -2.483 uVs 12:56:06 Integrated Voltage Measurement... 12:56:15 Voltage Offset Correction: v_off_ini = 2431.234 uV, v_off_fin = 2419.267 uV 12:56:15 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = 41.018 uVs 12:56:15 Integrated Voltage Measurement... 12:56:24 Voltage Offset Correction: v_off_ini = 2435.824 uV, v_off_fin = 2440.630 uV 12:56:24 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -54.840 uVs 12:56:24 Integrated Voltage Measurement... 12:56:33 Voltage Offset Correction: v_off_ini = 2434.174 uV, v_off_fin = 2419.957 uV 12:56:33 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = 55.021 uVs 12:56:37 I1X Average and RMS of Measurements VT = -40.260 +- 13.815 uVs I1X = 8.053 +- 2.763 uTm 12:56:40 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 12:56:44 Integrated Voltage Measurement... 12:56:53 Voltage Offset Correction: v_off_ini = 2414.588 uV, v_off_fin = 2418.877 uV 12:56:53 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -178.573 uVs 12:56:53 Integrated Voltage Measurement... 12:57:02 Voltage Offset Correction: v_off_ini = 2431.576 uV, v_off_fin = 2440.868 uV 12:57:02 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 210.141 uVs 12:57:02 Integrated Voltage Measurement... 12:57:11 Voltage Offset Correction: v_off_ini = 2423.526 uV, v_off_fin = 2403.492 uV 12:57:11 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -202.139 uVs 12:57:11 Integrated Voltage Measurement... 12:57:20 Voltage Offset Correction: v_off_ini = 2392.305 uV, v_off_fin = 2420.604 uV 12:57:20 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 202.423 uVs 12:57:20 Integrated Voltage Measurement... 12:57:29 Voltage Offset Correction: v_off_ini = 2401.435 uV, v_off_fin = 2411.124 uV 12:57:29 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -215.297 uVs 12:57:29 Integrated Voltage Measurement... 12:57:38 Voltage Offset Correction: v_off_ini = 2420.097 uV, v_off_fin = 2427.051 uV 12:57:38 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 171.736 uVs 12:57:42 I1Y Average and RMS of Measurements VT = -196.718 +- 3.948 uVs I1Y = -39.341 +- 0.789 uTm 12:57:47 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 12:57:50 Integrated Voltage Measurement... 12:57:59 Voltage Offset Correction: v_off_ini = 2398.898 uV, v_off_fin = 2393.892 uV 12:57:59 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 17.549 uVs 12:57:59 Integrated Voltage Measurement... 12:58:08 Voltage Offset Correction: v_off_ini = 2425.673 uV, v_off_fin = 2414.320 uV 12:58:08 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -40.465 uVs 12:58:08 Integrated Voltage Measurement... 12:58:17 Voltage Offset Correction: v_off_ini = 2402.793 uV, v_off_fin = 2394.669 uV 12:58:17 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 1.566 uVs 12:58:17 Integrated Voltage Measurement... 12:58:26 Voltage Offset Correction: v_off_ini = 2414.829 uV, v_off_fin = 2430.073 uV 12:58:26 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -19.843 uVs 12:58:26 Integrated Voltage Measurement... 12:58:35 Voltage Offset Correction: v_off_ini = 2417.795 uV, v_off_fin = 2414.928 uV 12:58:35 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 32.845 uVs 12:58:35 Integrated Voltage Measurement... 12:58:44 Voltage Offset Correction: v_off_ini = 2435.984 uV, v_off_fin = 2435.934 uV 12:58:44 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = -22.214 uVs 12:58:46 I2X Average and RMS of Measurements VT = 22.414 +- 8.301 uVs I2X = -2.062 +- 0.764 uTm^2 12:58:51 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 12:58:54 Integrated Voltage Measurement... 12:59:03 Voltage Offset Correction: v_off_ini = 2445.124 uV, v_off_fin = 2441.934 uV 12:59:03 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = -199.045 uVs 12:59:03 Integrated Voltage Measurement... 12:59:12 Voltage Offset Correction: v_off_ini = 2454.128 uV, v_off_fin = 2441.886 uV 12:59:12 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 213.457 uVs 12:59:12 Integrated Voltage Measurement... 12:59:21 Voltage Offset Correction: v_off_ini = 2466.098 uV, v_off_fin = 2446.183 uV 12:59:21 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -220.101 uVs 12:59:21 Integrated Voltage Measurement... 12:59:30 Voltage Offset Correction: v_off_ini = 2436.848 uV, v_off_fin = 2450.391 uV 12:59:30 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 189.743 uVs 12:59:30 Integrated Voltage Measurement... 12:59:39 Voltage Offset Correction: v_off_ini = 2459.521 uV, v_off_fin = 2439.870 uV 12:59:39 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -215.526 uVs 12:59:39 Integrated Voltage Measurement... 12:59:48 Voltage Offset Correction: v_off_ini = 2442.764 uV, v_off_fin = 2471.823 uV 12:59:48 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 227.846 uVs 12:59:50 I2Y Average and RMS of Measurements VT = -210.953 +- 7.609 uVs I2Y = -19.408 +- 0.700 uTm^2 13:00:02 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 13:00:07 Integrated Voltage Measurement... 13:00:15 Voltage Offset Correction: v_off_ini = 2480.316 uV, v_off_fin = 2467.628 uV 13:00:15 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -3.738 uVs 13:00:16 Integrated Voltage Measurement... 13:00:24 Voltage Offset Correction: v_off_ini = 2479.694 uV, v_off_fin = 2470.724 uV 13:00:24 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -8.221 uVs 13:00:25 Integrated Voltage Measurement... 13:00:33 Voltage Offset Correction: v_off_ini = 2444.353 uV, v_off_fin = 2458.536 uV 13:00:34 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -17.445 uVs 13:00:34 Integrated Voltage Measurement... 13:00:42 Voltage Offset Correction: v_off_ini = 2459.109 uV, v_off_fin = 2474.344 uV 13:00:43 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 25.685 uVs 13:00:43 Integrated Voltage Measurement... 13:00:51 Voltage Offset Correction: v_off_ini = 2468.752 uV, v_off_fin = 2485.489 uV 13:00:52 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -20.149 uVs 13:00:52 Integrated Voltage Measurement... 13:01:01 Voltage Offset Correction: v_off_ini = 2470.871 uV, v_off_fin = 2470.741 uV 13:01:01 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 19.896 uVs 13:01:05 I1X Average and RMS of Measurements VT = -13.115 +- 10.877 uVs I1X = 2.623 +- 2.175 uTm 13:01:07 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 13:01:11 Integrated Voltage Measurement... 13:01:20 Voltage Offset Correction: v_off_ini = 2481.511 uV, v_off_fin = 2477.660 uV 13:01:20 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = -193.208 uVs 13:01:20 Integrated Voltage Measurement... 13:01:29 Voltage Offset Correction: v_off_ini = 2500.899 uV, v_off_fin = 2510.472 uV 13:01:29 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 185.229 uVs 13:01:29 Integrated Voltage Measurement... 13:01:38 Voltage Offset Correction: v_off_ini = 2496.493 uV, v_off_fin = 2478.892 uV 13:01:38 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -205.736 uVs 13:01:38 Integrated Voltage Measurement... 13:01:47 Voltage Offset Correction: v_off_ini = 2484.862 uV, v_off_fin = 2496.508 uV 13:01:47 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 200.521 uVs 13:01:47 Integrated Voltage Measurement... 13:01:56 Voltage Offset Correction: v_off_ini = 2498.823 uV, v_off_fin = 2487.331 uV 13:01:56 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -203.328 uVs 13:01:56 Integrated Voltage Measurement... 13:02:05 Voltage Offset Correction: v_off_ini = 2489.856 uV, v_off_fin = 2504.713 uV 13:02:05 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 184.474 uVs 13:02:10 I1Y Average and RMS of Measurements VT = -195.416 +- 5.779 uVs I1Y = -39.081 +- 1.154 uTm 13:02:15 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 13:02:17 Integrated Voltage Measurement... 13:02:26 Voltage Offset Correction: v_off_ini = 2511.786 uV, v_off_fin = 2509.075 uV 13:02:26 Integrated Voltage Measurement dXYus = -1.0001 mm, dXYds = 0.0000 mm dXY = -1.0001 mm VT = 62.331 uVs 13:02:26 Integrated Voltage Measurement... 13:02:35 Voltage Offset Correction: v_off_ini = 2501.969 uV, v_off_fin = 2512.110 uV 13:02:35 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -51.635 uVs 13:02:35 Integrated Voltage Measurement... 13:02:44 Voltage Offset Correction: v_off_ini = 2515.051 uV, v_off_fin = 2497.434 uV 13:02:44 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 54.807 uVs 13:02:44 Integrated Voltage Measurement... 13:02:53 Voltage Offset Correction: v_off_ini = 2516.882 uV, v_off_fin = 2528.374 uV 13:02:53 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -50.118 uVs 13:02:53 Integrated Voltage Measurement... 13:03:02 Voltage Offset Correction: v_off_ini = 2510.131 uV, v_off_fin = 2499.708 uV 13:03:02 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 49.324 uVs 13:03:02 Integrated Voltage Measurement... 13:03:11 Voltage Offset Correction: v_off_ini = 2510.243 uV, v_off_fin = 2499.096 uV 13:03:11 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -53.629 uVs 13:03:14 I2X Average and RMS of Measurements VT = 53.641 +- 2.397 uVs I2X = -4.935 +- 0.220 uTm^2 13:03:19 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 13:03:21 Integrated Voltage Measurement... 13:03:30 Voltage Offset Correction: v_off_ini = 2474.606 uV, v_off_fin = 2485.609 uV 13:03:30 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -219.413 uVs 13:03:30 Integrated Voltage Measurement... 13:03:39 Voltage Offset Correction: v_off_ini = 2480.697 uV, v_off_fin = 2496.468 uV 13:03:39 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 221.009 uVs 13:03:39 Integrated Voltage Measurement... 13:03:48 Voltage Offset Correction: v_off_ini = 2519.989 uV, v_off_fin = 2502.348 uV 13:03:48 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -184.948 uVs 13:03:48 Integrated Voltage Measurement... 13:03:57 Voltage Offset Correction: v_off_ini = 2478.098 uV, v_off_fin = 2502.177 uV 13:03:57 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 213.854 uVs 13:03:58 Integrated Voltage Measurement... 13:04:06 Voltage Offset Correction: v_off_ini = 2494.834 uV, v_off_fin = 2498.664 uV 13:04:07 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -203.522 uVs 13:04:07 Integrated Voltage Measurement... 13:04:15 Voltage Offset Correction: v_off_ini = 2481.778 uV, v_off_fin = 2488.203 uV 13:04:16 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 189.696 uVs 13:04:18 I2Y Average and RMS of Measurements VT = -205.407 +- 10.530 uVs I2Y = -18.897 +- 0.969 uTm^2 13:04:30 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 13:04:34 Integrated Voltage Measurement... 13:04:43 Voltage Offset Correction: v_off_ini = 2493.630 uV, v_off_fin = 2479.578 uV 13:04:43 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 7.646 uVs 13:04:43 Integrated Voltage Measurement... 13:04:52 Voltage Offset Correction: v_off_ini = 2498.339 uV, v_off_fin = 2507.546 uV 13:04:52 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -34.462 uVs 13:04:52 Integrated Voltage Measurement... 13:05:01 Voltage Offset Correction: v_off_ini = 2522.584 uV, v_off_fin = 2501.211 uV 13:05:01 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -1.232 uVs 13:05:01 Integrated Voltage Measurement... 13:05:10 Voltage Offset Correction: v_off_ini = 2529.383 uV, v_off_fin = 2528.195 uV 13:05:10 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.4999 mm dXY = 0.4999 mm VT = -31.877 uVs 13:05:10 Integrated Voltage Measurement... 13:05:19 Voltage Offset Correction: v_off_ini = 2526.477 uV, v_off_fin = 2530.594 uV 13:05:19 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 18.271 uVs 13:05:19 Integrated Voltage Measurement... 13:05:28 Voltage Offset Correction: v_off_ini = 2520.609 uV, v_off_fin = 2524.871 uV 13:05:28 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -11.710 uVs 13:05:32 I1X Average and RMS of Measurements VT = 17.122 +- 2.783 uVs I1X = -3.425 +- 0.557 uTm 13:05:35 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 13:05:39 Integrated Voltage Measurement... 13:05:48 Voltage Offset Correction: v_off_ini = 2506.679 uV, v_off_fin = 2483.603 uV 13:05:48 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = -182.231 uVs 13:05:48 Integrated Voltage Measurement... 13:05:57 Voltage Offset Correction: v_off_ini = 2505.443 uV, v_off_fin = 2513.934 uV 13:05:57 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 207.585 uVs 13:05:57 Integrated Voltage Measurement... 13:06:06 Voltage Offset Correction: v_off_ini = 2497.700 uV, v_off_fin = 2492.838 uV 13:06:06 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -197.312 uVs 13:06:06 Integrated Voltage Measurement... 13:06:15 Voltage Offset Correction: v_off_ini = 2503.297 uV, v_off_fin = 2506.302 uV 13:06:15 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 188.837 uVs 13:06:15 Integrated Voltage Measurement... 13:06:24 Voltage Offset Correction: v_off_ini = 2485.943 uV, v_off_fin = 2491.495 uV 13:06:24 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -197.497 uVs 13:06:24 Integrated Voltage Measurement... 13:06:33 Voltage Offset Correction: v_off_ini = 2505.180 uV, v_off_fin = 2525.392 uV 13:06:33 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 175.269 uVs 13:06:37 I1Y Average and RMS of Measurements VT = -191.455 +- 3.664 uVs I1Y = -38.289 +- 0.733 uTm 13:06:42 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 13:06:45 Integrated Voltage Measurement... 13:06:54 Voltage Offset Correction: v_off_ini = 2500.188 uV, v_off_fin = 2476.775 uV 13:06:54 Integrated Voltage Measurement dXYus = -1.0001 mm, dXYds = 0.0000 mm dXY = -1.0001 mm VT = 72.587 uVs 13:06:54 Integrated Voltage Measurement... 13:07:03 Voltage Offset Correction: v_off_ini = 2501.676 uV, v_off_fin = 2502.563 uV 13:07:03 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -73.841 uVs 13:07:03 Integrated Voltage Measurement... 13:07:12 Voltage Offset Correction: v_off_ini = 2497.909 uV, v_off_fin = 2486.534 uV 13:07:12 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 86.035 uVs 13:07:12 Integrated Voltage Measurement... 13:07:21 Voltage Offset Correction: v_off_ini = 2469.060 uV, v_off_fin = 2481.120 uV 13:07:21 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -83.418 uVs 13:07:21 Integrated Voltage Measurement... 13:07:30 Voltage Offset Correction: v_off_ini = 2463.607 uV, v_off_fin = 2446.256 uV 13:07:30 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 92.910 uVs 13:07:30 Integrated Voltage Measurement... 13:07:39 Voltage Offset Correction: v_off_ini = 2473.332 uV, v_off_fin = 2476.722 uV 13:07:39 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -65.188 uVs 13:07:41 I2X Average and RMS of Measurements VT = 78.996 +- 4.700 uVs I2X = -7.268 +- 0.433 uTm^2 13:07:46 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 13:07:49 Integrated Voltage Measurement... 13:07:58 Voltage Offset Correction: v_off_ini = 2482.290 uV, v_off_fin = 2500.588 uV 13:07:58 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -231.150 uVs 13:07:58 Integrated Voltage Measurement... 13:08:07 Voltage Offset Correction: v_off_ini = 2471.556 uV, v_off_fin = 2509.698 uV 13:08:07 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 211.968 uVs 13:08:07 Integrated Voltage Measurement... 13:08:16 Voltage Offset Correction: v_off_ini = 2509.192 uV, v_off_fin = 2501.250 uV 13:08:16 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -218.832 uVs 13:08:16 Integrated Voltage Measurement... 13:08:25 Voltage Offset Correction: v_off_ini = 2505.085 uV, v_off_fin = 2520.191 uV 13:08:25 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 206.328 uVs 13:08:25 Integrated Voltage Measurement... 13:08:34 Voltage Offset Correction: v_off_ini = 2498.963 uV, v_off_fin = 2500.170 uV 13:08:34 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = -204.583 uVs 13:08:34 Integrated Voltage Measurement... 13:08:43 Voltage Offset Correction: v_off_ini = 2501.736 uV, v_off_fin = 2478.923 uV 13:08:43 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 178.518 uVs 13:08:45 I2Y Average and RMS of Measurements VT = -208.563 +- 12.576 uVs I2Y = -19.188 +- 1.157 uTm^2 13:09:00 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 13:09:04 Integrated Voltage Measurement... 13:09:13 Voltage Offset Correction: v_off_ini = 2520.586 uV, v_off_fin = 2501.345 uV 13:09:13 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 70.019 uVs 13:09:13 Integrated Voltage Measurement... 13:09:22 Voltage Offset Correction: v_off_ini = 2507.427 uV, v_off_fin = 2503.386 uV 13:09:22 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.4999 mm dXY = 0.4999 mm VT = -50.509 uVs 13:09:22 Integrated Voltage Measurement... 13:09:31 Voltage Offset Correction: v_off_ini = 2505.723 uV, v_off_fin = 2516.709 uV 13:09:31 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 52.027 uVs 13:09:31 Integrated Voltage Measurement... 13:09:40 Voltage Offset Correction: v_off_ini = 2522.717 uV, v_off_fin = 2511.816 uV 13:09:40 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -80.436 uVs 13:09:40 Integrated Voltage Measurement... 13:09:49 Voltage Offset Correction: v_off_ini = 2509.800 uV, v_off_fin = 2515.020 uV 13:09:49 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 66.610 uVs 13:09:49 Integrated Voltage Measurement... 13:09:58 Voltage Offset Correction: v_off_ini = 2511.604 uV, v_off_fin = 2514.448 uV 13:09:58 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -30.572 uVs 13:10:02 I1X Average and RMS of Measurements VT = 58.362 +- 7.326 uVs I1X = -11.673 +- 1.465 uTm 13:10:05 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 13:10:09 Integrated Voltage Measurement... 13:10:18 Voltage Offset Correction: v_off_ini = 2500.628 uV, v_off_fin = 2488.548 uV 13:10:18 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = -195.097 uVs 13:10:18 Integrated Voltage Measurement... 13:10:27 Voltage Offset Correction: v_off_ini = 2500.165 uV, v_off_fin = 2503.134 uV 13:10:27 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 174.060 uVs 13:10:27 Integrated Voltage Measurement... 13:10:36 Voltage Offset Correction: v_off_ini = 2521.522 uV, v_off_fin = 2491.454 uV 13:10:36 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -197.150 uVs 13:10:36 Integrated Voltage Measurement... 13:10:45 Voltage Offset Correction: v_off_ini = 2511.721 uV, v_off_fin = 2531.239 uV 13:10:45 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 183.232 uVs 13:10:45 Integrated Voltage Measurement... 13:10:54 Voltage Offset Correction: v_off_ini = 2510.750 uV, v_off_fin = 2500.120 uV 13:10:54 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -220.014 uVs 13:10:54 Integrated Voltage Measurement... 13:11:03 Voltage Offset Correction: v_off_ini = 2498.772 uV, v_off_fin = 2525.219 uV 13:11:03 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 173.329 uVs 13:11:07 I1Y Average and RMS of Measurements VT = -190.480 +- 4.941 uVs I1Y = -38.095 +- 0.987 uTm 13:11:12 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 13:11:15 Integrated Voltage Measurement... 13:11:24 Voltage Offset Correction: v_off_ini = 2488.511 uV, v_off_fin = 2516.705 uV 13:11:24 Integrated Voltage Measurement dXYus = -1.0001 mm, dXYds = 0.0000 mm dXY = -1.0001 mm VT = 107.321 uVs 13:11:24 Integrated Voltage Measurement... 13:11:33 Voltage Offset Correction: v_off_ini = 2524.424 uV, v_off_fin = 2519.117 uV 13:11:33 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -124.069 uVs 13:11:33 Integrated Voltage Measurement... 13:11:42 Voltage Offset Correction: v_off_ini = 2509.919 uV, v_off_fin = 2507.274 uV 13:11:42 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 139.986 uVs 13:11:42 Integrated Voltage Measurement... 13:11:51 Voltage Offset Correction: v_off_ini = 2530.363 uV, v_off_fin = 2521.945 uV 13:11:51 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -114.732 uVs 13:11:51 Integrated Voltage Measurement... 13:12:00 Voltage Offset Correction: v_off_ini = 2522.964 uV, v_off_fin = 2505.817 uV 13:12:00 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 111.035 uVs 13:12:00 Integrated Voltage Measurement... 13:12:09 Voltage Offset Correction: v_off_ini = 2539.601 uV, v_off_fin = 2540.755 uV 13:12:09 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -112.253 uVs 13:12:11 I2X Average and RMS of Measurements VT = 118.233 +- 6.662 uVs I2X = -10.877 +- 0.613 uTm^2 13:12:16 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 13:12:19 Integrated Voltage Measurement... 13:12:28 Voltage Offset Correction: v_off_ini = 2525.059 uV, v_off_fin = 2518.924 uV 13:12:28 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -212.019 uVs 13:12:28 Integrated Voltage Measurement... 13:12:37 Voltage Offset Correction: v_off_ini = 2527.081 uV, v_off_fin = 2531.443 uV 13:12:37 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 213.396 uVs 13:12:37 Integrated Voltage Measurement... 13:12:46 Voltage Offset Correction: v_off_ini = 2525.695 uV, v_off_fin = 2553.532 uV 13:12:46 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -228.432 uVs 13:12:46 Integrated Voltage Measurement... 13:12:55 Voltage Offset Correction: v_off_ini = 2555.260 uV, v_off_fin = 2533.998 uV 13:12:55 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 204.972 uVs 13:12:55 Integrated Voltage Measurement... 13:13:04 Voltage Offset Correction: v_off_ini = 2540.822 uV, v_off_fin = 2538.401 uV 13:13:04 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -214.404 uVs 13:13:04 Integrated Voltage Measurement... 13:13:13 Voltage Offset Correction: v_off_ini = 2554.299 uV, v_off_fin = 2568.587 uV 13:13:13 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 218.027 uVs 13:13:15 I2Y Average and RMS of Measurements VT = -215.208 +- 1.780 uVs I2Y = -19.799 +- 0.164 uTm^2 13:13:30 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 13:13:34 Integrated Voltage Measurement... 13:13:43 Voltage Offset Correction: v_off_ini = 2552.243 uV, v_off_fin = 2550.766 uV 13:13:43 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 63.820 uVs 13:13:43 Integrated Voltage Measurement... 13:13:52 Voltage Offset Correction: v_off_ini = 2538.240 uV, v_off_fin = 2541.527 uV 13:13:52 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -82.642 uVs 13:13:52 Integrated Voltage Measurement... 13:14:01 Voltage Offset Correction: v_off_ini = 2569.519 uV, v_off_fin = 2560.186 uV 13:14:01 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 93.685 uVs 13:14:01 Integrated Voltage Measurement... 13:14:10 Voltage Offset Correction: v_off_ini = 2558.586 uV, v_off_fin = 2566.754 uV 13:14:10 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -70.673 uVs 13:14:10 Integrated Voltage Measurement... 13:14:19 Voltage Offset Correction: v_off_ini = 2571.546 uV, v_off_fin = 2576.090 uV 13:14:19 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 74.284 uVs 13:14:19 Integrated Voltage Measurement... 13:14:28 Voltage Offset Correction: v_off_ini = 2594.156 uV, v_off_fin = 2558.213 uV 13:14:28 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -71.710 uVs 13:14:32 I1X Average and RMS of Measurements VT = 76.136 +- 4.274 uVs I1X = -15.227 +- 0.855 uTm 13:14:35 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 13:14:39 Integrated Voltage Measurement... 13:14:48 Voltage Offset Correction: v_off_ini = 2578.216 uV, v_off_fin = 2576.677 uV 13:14:48 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = -190.683 uVs 13:14:48 Integrated Voltage Measurement... 13:14:57 Voltage Offset Correction: v_off_ini = 2583.946 uV, v_off_fin = 2611.185 uV 13:14:57 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 215.352 uVs 13:14:57 Integrated Voltage Measurement... 13:15:06 Voltage Offset Correction: v_off_ini = 2575.395 uV, v_off_fin = 2568.905 uV 13:15:06 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -188.553 uVs 13:15:06 Integrated Voltage Measurement... 13:15:15 Voltage Offset Correction: v_off_ini = 2579.182 uV, v_off_fin = 2596.792 uV 13:15:15 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 208.213 uVs 13:15:15 Integrated Voltage Measurement... 13:15:24 Voltage Offset Correction: v_off_ini = 2589.965 uV, v_off_fin = 2568.491 uV 13:15:24 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -193.283 uVs 13:15:24 Integrated Voltage Measurement... 13:15:33 Voltage Offset Correction: v_off_ini = 2579.593 uV, v_off_fin = 2595.037 uV 13:15:33 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 201.578 uVs 13:15:37 I1Y Average and RMS of Measurements VT = -199.610 +- 2.440 uVs I1Y = -39.920 +- 0.489 uTm 13:15:42 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 13:15:45 Integrated Voltage Measurement... 13:15:54 Voltage Offset Correction: v_off_ini = 2569.316 uV, v_off_fin = 2562.417 uV 13:15:54 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 154.239 uVs 13:15:54 Integrated Voltage Measurement... 13:16:03 Voltage Offset Correction: v_off_ini = 2591.923 uV, v_off_fin = 2595.202 uV 13:16:03 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -144.361 uVs 13:16:03 Integrated Voltage Measurement... 13:16:12 Voltage Offset Correction: v_off_ini = 2599.894 uV, v_off_fin = 2587.207 uV 13:16:12 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 139.182 uVs 13:16:12 Integrated Voltage Measurement... 13:16:21 Voltage Offset Correction: v_off_ini = 2575.588 uV, v_off_fin = 2597.995 uV 13:16:21 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -115.969 uVs 13:16:21 Integrated Voltage Measurement... 13:16:30 Voltage Offset Correction: v_off_ini = 2612.847 uV, v_off_fin = 2592.164 uV 13:16:30 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 121.612 uVs 13:16:30 Integrated Voltage Measurement... 13:16:39 Voltage Offset Correction: v_off_ini = 2573.525 uV, v_off_fin = 2569.892 uV 13:16:39 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -107.079 uVs 13:16:41 I2X Average and RMS of Measurements VT = 130.407 +- 14.410 uVs I2X = -11.997 +- 1.326 uTm^2 13:16:46 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 13:16:49 Integrated Voltage Measurement... 13:16:58 Voltage Offset Correction: v_off_ini = 2597.169 uV, v_off_fin = 2591.073 uV 13:16:58 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -205.196 uVs 13:16:58 Integrated Voltage Measurement... 13:17:07 Voltage Offset Correction: v_off_ini = 2599.416 uV, v_off_fin = 2619.080 uV 13:17:07 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 217.911 uVs 13:17:07 Integrated Voltage Measurement... 13:17:16 Voltage Offset Correction: v_off_ini = 2607.273 uV, v_off_fin = 2622.288 uV 13:17:16 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -209.086 uVs 13:17:16 Integrated Voltage Measurement... 13:17:25 Voltage Offset Correction: v_off_ini = 2612.906 uV, v_off_fin = 2629.936 uV 13:17:25 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 222.873 uVs 13:17:25 Integrated Voltage Measurement... 13:17:34 Voltage Offset Correction: v_off_ini = 2621.659 uV, v_off_fin = 2621.113 uV 13:17:34 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -232.436 uVs 13:17:34 Integrated Voltage Measurement... 13:17:43 Voltage Offset Correction: v_off_ini = 2632.416 uV, v_off_fin = 2638.984 uV 13:17:43 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 194.035 uVs 13:17:45 I2Y Average and RMS of Measurements VT = -213.589 +- 1.824 uVs I2Y = -19.650 +- 0.168 uTm^2 13:17:59 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 13:18:04 Integrated Voltage Measurement... 13:18:13 Voltage Offset Correction: v_off_ini = 2628.399 uV, v_off_fin = 2627.702 uV 13:18:13 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 70.395 uVs 13:18:13 Integrated Voltage Measurement... 13:18:22 Voltage Offset Correction: v_off_ini = 2628.258 uV, v_off_fin = 2619.835 uV 13:18:22 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -95.912 uVs 13:18:22 Integrated Voltage Measurement... 13:18:31 Voltage Offset Correction: v_off_ini = 2622.031 uV, v_off_fin = 2617.696 uV 13:18:31 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = 92.570 uVs 13:18:31 Integrated Voltage Measurement... 13:18:40 Voltage Offset Correction: v_off_ini = 2610.548 uV, v_off_fin = 2610.073 uV 13:18:40 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -77.939 uVs 13:18:40 Integrated Voltage Measurement... 13:18:49 Voltage Offset Correction: v_off_ini = 2603.002 uV, v_off_fin = 2595.408 uV 13:18:49 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 82.196 uVs 13:18:49 Integrated Voltage Measurement... 13:18:58 Voltage Offset Correction: v_off_ini = 2606.327 uV, v_off_fin = 2609.697 uV 13:18:58 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.4999 mm dXY = 0.4999 mm VT = -90.235 uVs 13:19:02 I1X Average and RMS of Measurements VT = 84.874 +- 1.278 uVs I1X = -16.976 +- 0.257 uTm 13:19:04 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 13:19:09 Integrated Voltage Measurement... 13:19:18 Voltage Offset Correction: v_off_ini = 2629.631 uV, v_off_fin = 2627.127 uV 13:19:18 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = -186.393 uVs 13:19:18 Integrated Voltage Measurement... 13:19:27 Voltage Offset Correction: v_off_ini = 2641.282 uV, v_off_fin = 2652.062 uV 13:19:27 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 199.169 uVs 13:19:27 Integrated Voltage Measurement... 13:19:36 Voltage Offset Correction: v_off_ini = 2619.413 uV, v_off_fin = 2647.126 uV 13:19:36 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -180.783 uVs 13:19:36 Integrated Voltage Measurement... 13:19:45 Voltage Offset Correction: v_off_ini = 2668.501 uV, v_off_fin = 2683.691 uV 13:19:45 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 203.565 uVs 13:19:45 Integrated Voltage Measurement... 13:19:54 Voltage Offset Correction: v_off_ini = 2666.840 uV, v_off_fin = 2648.592 uV 13:19:54 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -205.655 uVs 13:19:54 Integrated Voltage Measurement... 13:20:03 Voltage Offset Correction: v_off_ini = 2649.183 uV, v_off_fin = 2664.108 uV 13:20:03 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 203.647 uVs 13:20:07 I1Y Average and RMS of Measurements VT = -196.535 +- 5.744 uVs I1Y = -39.304 +- 1.148 uTm 13:20:12 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 13:20:14 Integrated Voltage Measurement... 13:20:23 Voltage Offset Correction: v_off_ini = 2629.468 uV, v_off_fin = 2624.959 uV 13:20:23 Integrated Voltage Measurement dXYus = -1.0001 mm, dXYds = 0.0000 mm dXY = -1.0001 mm VT = 156.583 uVs 13:20:23 Integrated Voltage Measurement... 13:20:32 Voltage Offset Correction: v_off_ini = 2643.202 uV, v_off_fin = 2659.869 uV 13:20:32 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -166.974 uVs 13:20:32 Integrated Voltage Measurement... 13:20:41 Voltage Offset Correction: v_off_ini = 2655.141 uV, v_off_fin = 2642.626 uV 13:20:41 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 164.138 uVs 13:20:42 Integrated Voltage Measurement... 13:20:50 Voltage Offset Correction: v_off_ini = 2671.040 uV, v_off_fin = 2662.243 uV 13:20:51 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -150.512 uVs 13:20:51 Integrated Voltage Measurement... 13:21:00 Voltage Offset Correction: v_off_ini = 2650.808 uV, v_off_fin = 2648.381 uV 13:21:00 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 120.691 uVs 13:21:00 Integrated Voltage Measurement... 13:21:09 Voltage Offset Correction: v_off_ini = 2659.063 uV, v_off_fin = 2655.438 uV 13:21:09 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -157.170 uVs 13:21:11 I2X Average and RMS of Measurements VT = 152.678 +- 9.889 uVs I2X = -14.046 +- 0.910 uTm^2 13:21:16 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 13:21:19 Integrated Voltage Measurement... 13:21:28 Voltage Offset Correction: v_off_ini = 2627.230 uV, v_off_fin = 2631.267 uV 13:21:28 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -206.137 uVs 13:21:28 Integrated Voltage Measurement... 13:21:37 Voltage Offset Correction: v_off_ini = 2643.346 uV, v_off_fin = 2645.618 uV 13:21:37 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 208.635 uVs 13:21:37 Integrated Voltage Measurement... 13:21:46 Voltage Offset Correction: v_off_ini = 2611.920 uV, v_off_fin = 2624.662 uV 13:21:46 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -183.726 uVs 13:21:46 Integrated Voltage Measurement... 13:21:55 Voltage Offset Correction: v_off_ini = 2638.305 uV, v_off_fin = 2630.003 uV 13:21:55 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 244.453 uVs 13:21:55 Integrated Voltage Measurement... 13:22:04 Voltage Offset Correction: v_off_ini = 2619.741 uV, v_off_fin = 2618.038 uV 13:22:04 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -189.563 uVs 13:22:04 Integrated Voltage Measurement... 13:22:13 Voltage Offset Correction: v_off_ini = 2636.669 uV, v_off_fin = 2632.210 uV 13:22:13 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 202.800 uVs 13:22:15 I2Y Average and RMS of Measurements VT = -205.886 +- 7.388 uVs I2Y = -18.941 +- 0.680 uTm^2 13:22:29 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 13:22:34 Integrated Voltage Measurement... 13:22:42 Voltage Offset Correction: v_off_ini = 2645.152 uV, v_off_fin = 2662.276 uV 13:22:43 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 76.007 uVs 13:22:43 Integrated Voltage Measurement... 13:22:51 Voltage Offset Correction: v_off_ini = 2656.238 uV, v_off_fin = 2657.524 uV 13:22:51 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -85.704 uVs 13:22:52 Integrated Voltage Measurement... 13:23:00 Voltage Offset Correction: v_off_ini = 2622.429 uV, v_off_fin = 2627.596 uV 13:23:00 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = 44.514 uVs 13:23:01 Integrated Voltage Measurement... 13:23:09 Voltage Offset Correction: v_off_ini = 2633.816 uV, v_off_fin = 2636.471 uV 13:23:09 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -77.115 uVs 13:23:09 Integrated Voltage Measurement... 13:23:18 Voltage Offset Correction: v_off_ini = 2635.497 uV, v_off_fin = 2614.868 uV 13:23:18 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 92.857 uVs 13:23:18 Integrated Voltage Measurement... 13:23:27 Voltage Offset Correction: v_off_ini = 2607.925 uV, v_off_fin = 2627.301 uV 13:23:27 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -68.168 uVs 13:23:32 I1X Average and RMS of Measurements VT = 74.061 +- 9.368 uVs I1X = -14.812 +- 1.873 uTm 13:23:34 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 13:23:38 Integrated Voltage Measurement... 13:23:47 Voltage Offset Correction: v_off_ini = 2613.874 uV, v_off_fin = 2587.787 uV 13:23:47 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.4998 mm dXY = 0.4999 mm VT = -202.336 uVs 13:23:47 Integrated Voltage Measurement... 13:23:56 Voltage Offset Correction: v_off_ini = 2604.605 uV, v_off_fin = 2627.109 uV 13:23:56 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 197.843 uVs 13:23:56 Integrated Voltage Measurement... 13:24:05 Voltage Offset Correction: v_off_ini = 2618.919 uV, v_off_fin = 2601.533 uV 13:24:05 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -200.300 uVs 13:24:05 Integrated Voltage Measurement... 13:24:14 Voltage Offset Correction: v_off_ini = 2641.237 uV, v_off_fin = 2646.878 uV 13:24:14 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 186.305 uVs 13:24:14 Integrated Voltage Measurement... 13:24:23 Voltage Offset Correction: v_off_ini = 2644.348 uV, v_off_fin = 2637.343 uV 13:24:23 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -229.659 uVs 13:24:23 Integrated Voltage Measurement... 13:24:32 Voltage Offset Correction: v_off_ini = 2661.010 uV, v_off_fin = 2659.109 uV 13:24:32 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 204.221 uVs 13:24:36 I1Y Average and RMS of Measurements VT = -203.444 +- 9.937 uVs I1Y = -40.691 +- 1.987 uTm 13:24:41 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 13:24:44 Integrated Voltage Measurement... 13:24:53 Voltage Offset Correction: v_off_ini = 2627.870 uV, v_off_fin = 2630.212 uV 13:24:53 Integrated Voltage Measurement dXYus = -1.0001 mm, dXYds = 0.0000 mm dXY = -1.0001 mm VT = 151.172 uVs 13:24:53 Integrated Voltage Measurement... 13:25:02 Voltage Offset Correction: v_off_ini = 2619.002 uV, v_off_fin = 2610.190 uV 13:25:02 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -108.681 uVs 13:25:02 Integrated Voltage Measurement... 13:25:11 Voltage Offset Correction: v_off_ini = 2624.858 uV, v_off_fin = 2610.199 uV 13:25:11 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = 143.068 uVs 13:25:11 Integrated Voltage Measurement... 13:25:20 Voltage Offset Correction: v_off_ini = 2614.377 uV, v_off_fin = 2623.436 uV 13:25:20 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = -144.962 uVs 13:25:20 Integrated Voltage Measurement... 13:25:29 Voltage Offset Correction: v_off_ini = 2598.152 uV, v_off_fin = 2597.236 uV 13:25:29 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = 135.565 uVs 13:25:29 Integrated Voltage Measurement... 13:25:38 Voltage Offset Correction: v_off_ini = 2593.819 uV, v_off_fin = 2599.428 uV 13:25:38 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -129.765 uVs 13:25:40 I2X Average and RMS of Measurements VT = 135.536 +- 6.099 uVs I2X = -12.470 +- 0.562 uTm^2 13:25:46 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 13:25:48 Integrated Voltage Measurement... 13:25:57 Voltage Offset Correction: v_off_ini = 2578.880 uV, v_off_fin = 2579.265 uV 13:25:57 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -211.190 uVs 13:25:57 Integrated Voltage Measurement... 13:26:06 Voltage Offset Correction: v_off_ini = 2588.975 uV, v_off_fin = 2581.044 uV 13:26:06 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 199.585 uVs 13:26:06 Integrated Voltage Measurement... 13:26:15 Voltage Offset Correction: v_off_ini = 2609.637 uV, v_off_fin = 2586.124 uV 13:26:15 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -213.545 uVs 13:26:15 Integrated Voltage Measurement... 13:26:24 Voltage Offset Correction: v_off_ini = 2583.903 uV, v_off_fin = 2596.134 uV 13:26:24 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 198.486 uVs 13:26:24 Integrated Voltage Measurement... 13:26:33 Voltage Offset Correction: v_off_ini = 2581.713 uV, v_off_fin = 2570.993 uV 13:26:33 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -203.244 uVs 13:26:33 Integrated Voltage Measurement... 13:26:42 Voltage Offset Correction: v_off_ini = 2605.165 uV, v_off_fin = 2613.269 uV 13:26:42 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 238.805 uVs 13:26:45 I2Y Average and RMS of Measurements VT = -210.809 +- 7.228 uVs I2Y = -19.394 +- 0.665 uTm^2 13:26:59 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 13:27:03 Integrated Voltage Measurement... 13:27:12 Voltage Offset Correction: v_off_ini = 2597.497 uV, v_off_fin = 2596.402 uV 13:27:12 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = 66.439 uVs 13:27:12 Integrated Voltage Measurement... 13:27:21 Voltage Offset Correction: v_off_ini = 2587.282 uV, v_off_fin = 2630.753 uV 13:27:21 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = -63.131 uVs 13:27:21 Integrated Voltage Measurement... 13:27:30 Voltage Offset Correction: v_off_ini = 2605.150 uV, v_off_fin = 2617.475 uV 13:27:30 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 68.444 uVs 13:27:30 Integrated Voltage Measurement... 13:27:39 Voltage Offset Correction: v_off_ini = 2610.141 uV, v_off_fin = 2616.546 uV 13:27:39 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = -73.410 uVs 13:27:39 Integrated Voltage Measurement... 13:27:48 Voltage Offset Correction: v_off_ini = 2616.071 uV, v_off_fin = 2619.514 uV 13:27:48 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 33.719 uVs 13:27:48 Integrated Voltage Measurement... 13:27:57 Voltage Offset Correction: v_off_ini = 2646.009 uV, v_off_fin = 2631.769 uV 13:27:57 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -68.120 uVs 13:28:01 I1X Average and RMS of Measurements VT = 62.211 +- 8.368 uVs I1X = -12.443 +- 1.674 uTm 13:28:04 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 13:28:08 Integrated Voltage Measurement... 13:28:17 Voltage Offset Correction: v_off_ini = 2598.738 uV, v_off_fin = 2610.575 uV 13:28:17 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -200.191 uVs 13:28:17 Integrated Voltage Measurement... 13:28:26 Voltage Offset Correction: v_off_ini = 2613.069 uV, v_off_fin = 2631.938 uV 13:28:26 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 215.583 uVs 13:28:26 Integrated Voltage Measurement... 13:28:35 Voltage Offset Correction: v_off_ini = 2620.918 uV, v_off_fin = 2615.319 uV 13:28:35 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -219.574 uVs 13:28:35 Integrated Voltage Measurement... 13:28:44 Voltage Offset Correction: v_off_ini = 2616.959 uV, v_off_fin = 2644.712 uV 13:28:44 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 207.424 uVs 13:28:44 Integrated Voltage Measurement... 13:28:53 Voltage Offset Correction: v_off_ini = 2609.130 uV, v_off_fin = 2612.157 uV 13:28:53 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -208.779 uVs 13:28:53 Integrated Voltage Measurement... 13:29:02 Voltage Offset Correction: v_off_ini = 2618.369 uV, v_off_fin = 2635.846 uV 13:29:02 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 189.155 uVs 13:29:06 I1Y Average and RMS of Measurements VT = -206.784 +- 5.984 uVs I1Y = -41.354 +- 1.197 uTm 13:29:11 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 13:29:13 Integrated Voltage Measurement... 13:29:22 Voltage Offset Correction: v_off_ini = 2590.557 uV, v_off_fin = 2600.968 uV 13:29:22 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 119.025 uVs 13:29:22 Integrated Voltage Measurement... 13:29:31 Voltage Offset Correction: v_off_ini = 2600.002 uV, v_off_fin = 2615.113 uV 13:29:31 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -134.026 uVs 13:29:31 Integrated Voltage Measurement... 13:29:40 Voltage Offset Correction: v_off_ini = 2608.751 uV, v_off_fin = 2592.577 uV 13:29:40 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 110.350 uVs 13:29:40 Integrated Voltage Measurement... 13:29:49 Voltage Offset Correction: v_off_ini = 2602.826 uV, v_off_fin = 2621.110 uV 13:29:49 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -126.020 uVs 13:29:49 Integrated Voltage Measurement... 13:29:58 Voltage Offset Correction: v_off_ini = 2645.359 uV, v_off_fin = 2597.937 uV 13:29:58 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 120.530 uVs 13:29:58 Integrated Voltage Measurement... 13:30:07 Voltage Offset Correction: v_off_ini = 2613.299 uV, v_off_fin = 2626.574 uV 13:30:07 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -118.091 uVs 13:30:10 I2X Average and RMS of Measurements VT = 121.341 +- 3.695 uVs I2X = -11.163 +- 0.340 uTm^2 13:30:15 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 13:30:17 Integrated Voltage Measurement... 13:30:26 Voltage Offset Correction: v_off_ini = 2621.346 uV, v_off_fin = 2605.006 uV 13:30:26 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -208.293 uVs 13:30:27 Integrated Voltage Measurement... 13:30:35 Voltage Offset Correction: v_off_ini = 2581.554 uV, v_off_fin = 2592.622 uV 13:30:35 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 236.421 uVs 13:30:36 Integrated Voltage Measurement... 13:30:44 Voltage Offset Correction: v_off_ini = 2596.519 uV, v_off_fin = 2606.341 uV 13:30:45 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -215.481 uVs 13:30:45 Integrated Voltage Measurement... 13:30:53 Voltage Offset Correction: v_off_ini = 2604.175 uV, v_off_fin = 2589.693 uV 13:30:54 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 204.691 uVs 13:30:54 Integrated Voltage Measurement... 13:31:03 Voltage Offset Correction: v_off_ini = 2586.923 uV, v_off_fin = 2602.968 uV 13:31:03 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -228.446 uVs 13:31:03 Integrated Voltage Measurement... 13:31:12 Voltage Offset Correction: v_off_ini = 2588.156 uV, v_off_fin = 2602.766 uV 13:31:12 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 225.769 uVs 13:31:14 I2Y Average and RMS of Measurements VT = -219.850 +- 7.171 uVs I2Y = -20.226 +- 0.660 uTm^2 13:31:28 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 13:31:32 Integrated Voltage Measurement... 13:31:41 Voltage Offset Correction: v_off_ini = 2634.732 uV, v_off_fin = 2638.660 uV 13:31:41 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 84.985 uVs 13:31:41 Integrated Voltage Measurement... 13:31:50 Voltage Offset Correction: v_off_ini = 2621.072 uV, v_off_fin = 2626.425 uV 13:31:50 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.4999 mm dXY = 0.4999 mm VT = -59.769 uVs 13:31:50 Integrated Voltage Measurement... 13:31:59 Voltage Offset Correction: v_off_ini = 2604.300 uV, v_off_fin = 2620.603 uV 13:31:59 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 83.008 uVs 13:31:59 Integrated Voltage Measurement... 13:32:08 Voltage Offset Correction: v_off_ini = 2635.451 uV, v_off_fin = 2616.837 uV 13:32:08 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -81.218 uVs 13:32:08 Integrated Voltage Measurement... 13:32:17 Voltage Offset Correction: v_off_ini = 2603.117 uV, v_off_fin = 2601.907 uV 13:32:17 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 59.720 uVs 13:32:17 Integrated Voltage Measurement... 13:32:26 Voltage Offset Correction: v_off_ini = 2605.773 uV, v_off_fin = 2620.007 uV 13:32:26 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -30.652 uVs 13:32:30 I1X Average and RMS of Measurements VT = 66.559 +- 15.627 uVs I1X = -13.312 +- 3.126 uTm 13:32:33 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 13:32:37 Integrated Voltage Measurement... 13:32:46 Voltage Offset Correction: v_off_ini = 2605.430 uV, v_off_fin = 2605.537 uV 13:32:46 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -192.800 uVs 13:32:46 Integrated Voltage Measurement... 13:32:55 Voltage Offset Correction: v_off_ini = 2618.008 uV, v_off_fin = 2629.245 uV 13:32:55 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 220.770 uVs 13:32:55 Integrated Voltage Measurement... 13:33:04 Voltage Offset Correction: v_off_ini = 2640.870 uV, v_off_fin = 2614.658 uV 13:33:04 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -230.197 uVs 13:33:04 Integrated Voltage Measurement... 13:33:13 Voltage Offset Correction: v_off_ini = 2612.125 uV, v_off_fin = 2623.514 uV 13:33:13 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 195.401 uVs 13:33:13 Integrated Voltage Measurement... 13:33:22 Voltage Offset Correction: v_off_ini = 2620.719 uV, v_off_fin = 2596.093 uV 13:33:22 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -213.835 uVs 13:33:22 Integrated Voltage Measurement... 13:33:31 Voltage Offset Correction: v_off_ini = 2610.548 uV, v_off_fin = 2607.234 uV 13:33:31 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 218.339 uVs 13:33:35 I1Y Average and RMS of Measurements VT = -211.890 +- 3.851 uVs I1Y = -42.377 +- 0.769 uTm 13:33:40 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 13:33:43 Integrated Voltage Measurement... 13:33:52 Voltage Offset Correction: v_off_ini = 2617.476 uV, v_off_fin = 2604.281 uV 13:33:52 Integrated Voltage Measurement dXYus = -1.0002 mm, dXYds = 0.0000 mm dXY = -1.0002 mm VT = 101.777 uVs 13:33:52 Integrated Voltage Measurement... 13:34:01 Voltage Offset Correction: v_off_ini = 2625.690 uV, v_off_fin = 2639.472 uV 13:34:01 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -105.152 uVs 13:34:01 Integrated Voltage Measurement... 13:34:10 Voltage Offset Correction: v_off_ini = 2601.007 uV, v_off_fin = 2595.615 uV 13:34:10 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 106.563 uVs 13:34:10 Integrated Voltage Measurement... 13:34:19 Voltage Offset Correction: v_off_ini = 2611.741 uV, v_off_fin = 2595.981 uV 13:34:19 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -102.930 uVs 13:34:19 Integrated Voltage Measurement... 13:34:28 Voltage Offset Correction: v_off_ini = 2619.232 uV, v_off_fin = 2596.756 uV 13:34:28 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 101.941 uVs 13:34:28 Integrated Voltage Measurement... 13:34:37 Voltage Offset Correction: v_off_ini = 2621.819 uV, v_off_fin = 2642.927 uV 13:34:37 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -109.741 uVs 13:34:39 I2X Average and RMS of Measurements VT = 104.684 +- 0.971 uVs I2X = -9.631 +- 0.090 uTm^2 13:34:44 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 13:34:47 Integrated Voltage Measurement... 13:34:56 Voltage Offset Correction: v_off_ini = 2629.538 uV, v_off_fin = 2633.504 uV 13:34:56 Integrated Voltage Measurement dXYus = 0.9998 mm, dXYds = 0.0000 mm dXY = 0.9998 mm VT = -226.085 uVs 13:34:56 Integrated Voltage Measurement... 13:35:05 Voltage Offset Correction: v_off_ini = 2649.516 uV, v_off_fin = 2634.142 uV 13:35:05 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 217.073 uVs 13:35:05 Integrated Voltage Measurement... 13:35:14 Voltage Offset Correction: v_off_ini = 2628.756 uV, v_off_fin = 2623.090 uV 13:35:14 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = -211.061 uVs 13:35:14 Integrated Voltage Measurement... 13:35:23 Voltage Offset Correction: v_off_ini = 2644.492 uV, v_off_fin = 2643.397 uV 13:35:23 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 233.546 uVs 13:35:23 Integrated Voltage Measurement... 13:35:32 Voltage Offset Correction: v_off_ini = 2643.424 uV, v_off_fin = 2647.534 uV 13:35:32 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -218.074 uVs 13:35:32 Integrated Voltage Measurement... 13:35:41 Voltage Offset Correction: v_off_ini = 2661.512 uV, v_off_fin = 2644.765 uV 13:35:41 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 210.835 uVs 13:35:43 I2Y Average and RMS of Measurements VT = -219.446 +- 3.542 uVs I2Y = -20.190 +- 0.327 uTm^2 13:35:46 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 14:21:53 Wire Position: X0 = 50.0000 mm, Y0 = 0.0000 mm