SLAC Magnetic Measurements Date: 12-04-2025 Time: 10:15:38 Project: LCLS-II-HE Magnet Type: Phase Shifter Magnet Name: HE_SXPS_15081 Dataset Number: 1 Run Purpose: Tuning Run Type: Run Number: 1 Measurement Device: Stretched Wire Operator: YL Comment: 10:16:37 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:16:39 Motion Parameters, X Controller, Axis 1 acc = 0.001000 m/s/s dcl = 0.001000 m/s/s vel = 0.001000 m/s pos = 0.000000 m kp = 188.600000 ki = 820.000000 kd = 491.400000 db = 0.000000 10:16:41 Motion Parameters, X Controller, Axis 2 acc = 0.001000 m/s/s dcl = 0.001000 m/s/s vel = 0.001000 m/s pos = 0.000000 m kp = 188.600000 ki = 820.000000 kd = 491.400000 db = 0.000000 10:16:42 Motion Parameters, Y Controller, Axis 1 acc = 0.001000 m/s/s dcl = 0.001000 m/s/s vel = 0.001000 m/s pos = 0.000000 m kp = 188.600000 ki = 820.000000 kd = 491.400000 db = 0.000000 10:16:44 Motion Parameters, Y Controller, Axis 2 acc = 0.001000 m/s/s dcl = 0.001000 m/s/s vel = 0.001000 m/s pos = 0.000000 m kp = 188.600000 ki = 820.000000 kd = 491.400000 db = 0.000000 10:18:16 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:18:20 Integrated Voltage Measurement... 10:18:29 Voltage Offset Correction: v_off_ini = -117.609 uV, v_off_fin = -117.477 uV 10:18:29 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -91.429 uVs 10:18:29 Integrated Voltage Measurement... 10:18:38 Voltage Offset Correction: v_off_ini = -102.607 uV, v_off_fin = -114.227 uV 10:18:38 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 54.161 uVs 10:18:38 Integrated Voltage Measurement... 10:18:47 Voltage Offset Correction: v_off_ini = -130.250 uV, v_off_fin = -135.422 uV 10:18:47 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -80.794 uVs 10:18:47 Integrated Voltage Measurement... 10:18:56 Voltage Offset Correction: v_off_ini = -137.452 uV, v_off_fin = -123.645 uV 10:18:56 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 87.767 uVs 10:18:56 Integrated Voltage Measurement... 10:19:05 Voltage Offset Correction: v_off_ini = -140.473 uV, v_off_fin = -140.922 uV 10:19:05 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -73.992 uVs 10:19:05 Integrated Voltage Measurement... 10:19:14 Voltage Offset Correction: v_off_ini = -154.589 uV, v_off_fin = -139.936 uV 10:19:14 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 81.548 uVs 10:19:18 I1X Average and RMS of Measurements VT = -78.282 +- 4.703 uVs I1X = 15.657 +- 0.941 uTm 10:19:21 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:19:25 Integrated Voltage Measurement... 10:19:34 Voltage Offset Correction: v_off_ini = -146.900 uV, v_off_fin = -159.821 uV 10:19:34 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -324.552 uVs 10:19:34 Integrated Voltage Measurement... 10:19:43 Voltage Offset Correction: v_off_ini = -141.179 uV, v_off_fin = -136.483 uV 10:19:43 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 291.072 uVs 10:19:43 Integrated Voltage Measurement... 10:19:52 Voltage Offset Correction: v_off_ini = -125.057 uV, v_off_fin = -150.651 uV 10:19:52 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -322.678 uVs 10:19:52 Integrated Voltage Measurement... 10:20:01 Voltage Offset Correction: v_off_ini = -145.218 uV, v_off_fin = -143.912 uV 10:20:01 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 315.573 uVs 10:20:01 Integrated Voltage Measurement... 10:20:10 Voltage Offset Correction: v_off_ini = -152.177 uV, v_off_fin = -147.048 uV 10:20:10 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -305.868 uVs 10:20:10 Integrated Voltage Measurement... 10:20:19 Voltage Offset Correction: v_off_ini = -136.673 uV, v_off_fin = -109.898 uV 10:20:19 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 308.774 uVs 10:20:23 I1Y Average and RMS of Measurements VT = -311.419 +- 5.453 uVs I1Y = -62.284 +- 1.091 uTm 10:20:28 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:20:31 Integrated Voltage Measurement... 10:20:40 Voltage Offset Correction: v_off_ini = -138.471 uV, v_off_fin = -147.430 uV 10:20:40 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -36.293 uVs 10:20:40 Integrated Voltage Measurement... 10:20:49 Voltage Offset Correction: v_off_ini = -136.385 uV, v_off_fin = -127.623 uV 10:20:49 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 30.999 uVs 10:20:49 Integrated Voltage Measurement... 10:20:58 Voltage Offset Correction: v_off_ini = -138.246 uV, v_off_fin = -155.453 uV 10:20:58 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -44.026 uVs 10:20:58 Integrated Voltage Measurement... 10:21:07 Voltage Offset Correction: v_off_ini = -147.264 uV, v_off_fin = -134.881 uV 10:21:07 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 71.463 uVs 10:21:07 Integrated Voltage Measurement... 10:21:16 Voltage Offset Correction: v_off_ini = -154.047 uV, v_off_fin = -156.882 uV 10:21:16 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -45.872 uVs 10:21:16 Integrated Voltage Measurement... 10:21:25 Voltage Offset Correction: v_off_ini = -148.968 uV, v_off_fin = -137.892 uV 10:21:25 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 20.306 uVs 10:21:27 I2X Average and RMS of Measurements VT = -41.493 +- 11.494 uVs I2X = 3.817 +- 1.057 uTm^2 10:21:32 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:21:35 Integrated Voltage Measurement... 10:21:44 Voltage Offset Correction: v_off_ini = -134.193 uV, v_off_fin = -140.202 uV 10:21:44 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -321.961 uVs 10:21:44 Integrated Voltage Measurement... 10:21:53 Voltage Offset Correction: v_off_ini = -135.861 uV, v_off_fin = -129.886 uV 10:21:53 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 301.403 uVs 10:21:53 Integrated Voltage Measurement... 10:22:02 Voltage Offset Correction: v_off_ini = -158.679 uV, v_off_fin = -137.318 uV 10:22:02 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -291.501 uVs 10:22:02 Integrated Voltage Measurement... 10:22:11 Voltage Offset Correction: v_off_ini = -144.275 uV, v_off_fin = -140.441 uV 10:22:11 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 317.718 uVs 10:22:11 Integrated Voltage Measurement... 10:22:20 Voltage Offset Correction: v_off_ini = -123.555 uV, v_off_fin = -131.022 uV 10:22:20 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -318.987 uVs 10:22:20 Integrated Voltage Measurement... 10:22:29 Voltage Offset Correction: v_off_ini = -138.532 uV, v_off_fin = -149.902 uV 10:22:29 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 308.211 uVs 10:22:31 I2Y Average and RMS of Measurements VT = -309.964 +- 3.866 uVs I2Y = -28.517 +- 0.356 uTm^2 10:22:42 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:22:46 Integrated Voltage Measurement... 10:22:55 Voltage Offset Correction: v_off_ini = -129.507 uV, v_off_fin = -118.532 uV 10:22:55 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -85.840 uVs 10:22:55 Integrated Voltage Measurement... 10:23:04 Voltage Offset Correction: v_off_ini = -124.274 uV, v_off_fin = -118.569 uV 10:23:04 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 90.915 uVs 10:23:04 Integrated Voltage Measurement... 10:23:13 Voltage Offset Correction: v_off_ini = -129.273 uV, v_off_fin = -125.675 uV 10:23:13 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -70.101 uVs 10:23:13 Integrated Voltage Measurement... 10:23:22 Voltage Offset Correction: v_off_ini = -125.027 uV, v_off_fin = -125.748 uV 10:23:22 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 77.228 uVs 10:23:22 Integrated Voltage Measurement... 10:23:31 Voltage Offset Correction: v_off_ini = -135.510 uV, v_off_fin = -127.394 uV 10:23:31 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -64.248 uVs 10:23:31 Integrated Voltage Measurement... 10:23:40 Voltage Offset Correction: v_off_ini = -147.876 uV, v_off_fin = -128.329 uV 10:23:40 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 84.470 uVs 10:23:44 I1X Average and RMS of Measurements VT = -78.800 +- 6.778 uVs I1X = 15.760 +- 1.356 uTm 10:23:47 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:23:51 Integrated Voltage Measurement... 10:24:00 Voltage Offset Correction: v_off_ini = -119.295 uV, v_off_fin = -140.305 uV 10:24:00 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -306.870 uVs 10:24:00 Integrated Voltage Measurement... 10:24:09 Voltage Offset Correction: v_off_ini = -119.914 uV, v_off_fin = -119.381 uV 10:24:09 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 293.617 uVs 10:24:09 Integrated Voltage Measurement... 10:24:18 Voltage Offset Correction: v_off_ini = -116.023 uV, v_off_fin = -138.735 uV 10:24:18 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -310.753 uVs 10:24:18 Integrated Voltage Measurement... 10:24:27 Voltage Offset Correction: v_off_ini = -115.779 uV, v_off_fin = -109.816 uV 10:24:27 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 287.810 uVs 10:24:27 Integrated Voltage Measurement... 10:24:36 Voltage Offset Correction: v_off_ini = -126.269 uV, v_off_fin = -141.671 uV 10:24:36 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -300.784 uVs 10:24:36 Integrated Voltage Measurement... 10:24:45 Voltage Offset Correction: v_off_ini = -124.590 uV, v_off_fin = -108.902 uV 10:24:45 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 301.559 uVs 10:24:49 I1Y Average and RMS of Measurements VT = -300.232 +- 0.772 uVs I1Y = -60.046 +- 0.154 uTm 10:24:54 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:24:56 Integrated Voltage Measurement... 10:25:05 Voltage Offset Correction: v_off_ini = -130.359 uV, v_off_fin = -163.827 uV 10:25:05 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -52.036 uVs 10:25:05 Integrated Voltage Measurement... 10:25:14 Voltage Offset Correction: v_off_ini = -123.818 uV, v_off_fin = -113.280 uV 10:25:14 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 55.085 uVs 10:25:14 Integrated Voltage Measurement... 10:25:23 Voltage Offset Correction: v_off_ini = -133.341 uV, v_off_fin = -128.816 uV 10:25:23 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -36.356 uVs 10:25:23 Integrated Voltage Measurement... 10:25:32 Voltage Offset Correction: v_off_ini = -121.623 uV, v_off_fin = -115.861 uV 10:25:32 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 21.829 uVs 10:25:32 Integrated Voltage Measurement... 10:25:41 Voltage Offset Correction: v_off_ini = -109.322 uV, v_off_fin = -137.085 uV 10:25:41 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -42.701 uVs 10:25:41 Integrated Voltage Measurement... 10:25:50 Voltage Offset Correction: v_off_ini = -104.577 uV, v_off_fin = -109.593 uV 10:25:50 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 33.679 uVs 10:25:53 I2X Average and RMS of Measurements VT = -40.281 +- 10.098 uVs I2X = 3.706 +- 0.929 uTm^2 10:25:58 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:26:00 Integrated Voltage Measurement... 10:26:09 Voltage Offset Correction: v_off_ini = -141.154 uV, v_off_fin = -134.947 uV 10:26:09 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -298.749 uVs 10:26:09 Integrated Voltage Measurement... 10:26:18 Voltage Offset Correction: v_off_ini = -126.388 uV, v_off_fin = -138.154 uV 10:26:18 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 311.363 uVs 10:26:19 Integrated Voltage Measurement... 10:26:27 Voltage Offset Correction: v_off_ini = -138.724 uV, v_off_fin = -139.323 uV 10:26:27 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -271.947 uVs 10:26:28 Integrated Voltage Measurement... 10:26:36 Voltage Offset Correction: v_off_ini = -120.246 uV, v_off_fin = -136.212 uV 10:26:36 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 282.985 uVs 10:26:37 Integrated Voltage Measurement... 10:26:45 Voltage Offset Correction: v_off_ini = -134.454 uV, v_off_fin = -141.841 uV 10:26:46 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -304.130 uVs 10:26:46 Integrated Voltage Measurement... 10:26:54 Voltage Offset Correction: v_off_ini = -128.256 uV, v_off_fin = -137.010 uV 10:26:55 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 292.990 uVs 10:26:57 I2Y Average and RMS of Measurements VT = -293.694 +- 11.778 uVs I2Y = -27.020 +- 1.084 uTm^2 10:27:08 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:27:12 Integrated Voltage Measurement... 10:27:21 Voltage Offset Correction: v_off_ini = -160.635 uV, v_off_fin = -160.834 uV 10:27:21 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -46.901 uVs 10:27:21 Integrated Voltage Measurement... 10:27:30 Voltage Offset Correction: v_off_ini = -140.344 uV, v_off_fin = -135.320 uV 10:27:30 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 41.666 uVs 10:27:30 Integrated Voltage Measurement... 10:27:39 Voltage Offset Correction: v_off_ini = -163.135 uV, v_off_fin = -167.942 uV 10:27:39 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -33.204 uVs 10:27:39 Integrated Voltage Measurement... 10:27:48 Voltage Offset Correction: v_off_ini = -152.607 uV, v_off_fin = -145.807 uV 10:27:48 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 47.200 uVs 10:27:48 Integrated Voltage Measurement... 10:27:57 Voltage Offset Correction: v_off_ini = -153.915 uV, v_off_fin = -171.303 uV 10:27:57 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -55.195 uVs 10:27:57 Integrated Voltage Measurement... 10:28:06 Voltage Offset Correction: v_off_ini = -173.846 uV, v_off_fin = -158.968 uV 10:28:06 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 55.730 uVs 10:28:10 I1X Average and RMS of Measurements VT = -46.649 +- 6.451 uVs I1X = 9.330 +- 1.290 uTm 10:28:13 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:28:17 Integrated Voltage Measurement... 10:28:26 Voltage Offset Correction: v_off_ini = -173.698 uV, v_off_fin = -174.888 uV 10:28:26 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = -284.849 uVs 10:28:26 Integrated Voltage Measurement... 10:28:35 Voltage Offset Correction: v_off_ini = -154.199 uV, v_off_fin = -148.419 uV 10:28:35 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 281.590 uVs 10:28:35 Integrated Voltage Measurement... 10:28:44 Voltage Offset Correction: v_off_ini = -167.493 uV, v_off_fin = -166.255 uV 10:28:44 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -275.205 uVs 10:28:44 Integrated Voltage Measurement... 10:28:53 Voltage Offset Correction: v_off_ini = -162.904 uV, v_off_fin = -134.880 uV 10:28:53 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 292.963 uVs 10:28:53 Integrated Voltage Measurement... 10:29:02 Voltage Offset Correction: v_off_ini = -152.361 uV, v_off_fin = -165.858 uV 10:29:02 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -276.463 uVs 10:29:02 Integrated Voltage Measurement... 10:29:11 Voltage Offset Correction: v_off_ini = -176.357 uV, v_off_fin = -149.236 uV 10:29:11 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 292.541 uVs 10:29:15 I1Y Average and RMS of Measurements VT = -283.935 +- 0.534 uVs I1Y = -56.787 +- 0.105 uTm 10:29:20 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:29:23 Integrated Voltage Measurement... 10:29:32 Voltage Offset Correction: v_off_ini = -184.987 uV, v_off_fin = -171.530 uV 10:29:32 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -34.038 uVs 10:29:32 Integrated Voltage Measurement... 10:29:41 Voltage Offset Correction: v_off_ini = -163.180 uV, v_off_fin = -170.818 uV 10:29:41 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 42.093 uVs 10:29:41 Integrated Voltage Measurement... 10:29:50 Voltage Offset Correction: v_off_ini = -184.048 uV, v_off_fin = -184.434 uV 10:29:50 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -16.814 uVs 10:29:50 Integrated Voltage Measurement... 10:29:59 Voltage Offset Correction: v_off_ini = -176.902 uV, v_off_fin = -172.932 uV 10:29:59 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 21.154 uVs 10:29:59 Integrated Voltage Measurement... 10:30:08 Voltage Offset Correction: v_off_ini = -175.081 uV, v_off_fin = -178.791 uV 10:30:08 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -23.847 uVs 10:30:08 Integrated Voltage Measurement... 10:30:17 Voltage Offset Correction: v_off_ini = -171.398 uV, v_off_fin = -163.797 uV 10:30:17 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 42.024 uVs 10:30:19 I2X Average and RMS of Measurements VT = -29.995 +- 8.062 uVs I2X = 2.760 +- 0.742 uTm^2 10:30:24 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:30:27 Integrated Voltage Measurement... 10:30:36 Voltage Offset Correction: v_off_ini = -168.387 uV, v_off_fin = -174.616 uV 10:30:36 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -315.484 uVs 10:30:36 Integrated Voltage Measurement... 10:30:45 Voltage Offset Correction: v_off_ini = -200.826 uV, v_off_fin = -198.037 uV 10:30:45 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 317.197 uVs 10:30:45 Integrated Voltage Measurement... 10:30:54 Voltage Offset Correction: v_off_ini = -176.648 uV, v_off_fin = -177.032 uV 10:30:54 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -276.616 uVs 10:30:54 Integrated Voltage Measurement... 10:31:03 Voltage Offset Correction: v_off_ini = -182.820 uV, v_off_fin = -169.109 uV 10:31:03 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 297.916 uVs 10:31:03 Integrated Voltage Measurement... 10:31:12 Voltage Offset Correction: v_off_ini = -179.168 uV, v_off_fin = -185.702 uV 10:31:12 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -289.933 uVs 10:31:12 Integrated Voltage Measurement... 10:31:21 Voltage Offset Correction: v_off_ini = -185.528 uV, v_off_fin = -166.598 uV 10:31:21 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 305.054 uVs 10:31:24 I2Y Average and RMS of Measurements VT = -300.367 +- 12.042 uVs I2Y = -27.634 +- 1.108 uTm^2 10:31:34 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:31:39 Integrated Voltage Measurement... 10:31:48 Voltage Offset Correction: v_off_ini = -152.240 uV, v_off_fin = -162.877 uV 10:31:48 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -72.675 uVs 10:31:48 Integrated Voltage Measurement... 10:31:57 Voltage Offset Correction: v_off_ini = -155.993 uV, v_off_fin = -165.440 uV 10:31:57 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 62.066 uVs 10:31:57 Integrated Voltage Measurement... 10:32:06 Voltage Offset Correction: v_off_ini = -169.238 uV, v_off_fin = -163.849 uV 10:32:06 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -16.039 uVs 10:32:06 Integrated Voltage Measurement... 10:32:15 Voltage Offset Correction: v_off_ini = -192.256 uV, v_off_fin = -167.864 uV 10:32:15 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 73.058 uVs 10:32:15 Integrated Voltage Measurement... 10:32:24 Voltage Offset Correction: v_off_ini = -187.989 uV, v_off_fin = -169.021 uV 10:32:24 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -67.103 uVs 10:32:24 Integrated Voltage Measurement... 10:32:33 Voltage Offset Correction: v_off_ini = -179.324 uV, v_off_fin = -169.473 uV 10:32:33 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 98.975 uVs 10:32:37 I1X Average and RMS of Measurements VT = -64.986 +- 15.804 uVs I1X = 12.997 +- 3.161 uTm 10:32:40 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:32:44 Integrated Voltage Measurement... 10:32:53 Voltage Offset Correction: v_off_ini = -180.863 uV, v_off_fin = -166.814 uV 10:32:53 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4998 mm dXY = 0.4999 mm VT = -288.729 uVs 10:32:53 Integrated Voltage Measurement... 10:33:02 Voltage Offset Correction: v_off_ini = -150.533 uV, v_off_fin = -128.903 uV 10:33:02 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 272.941 uVs 10:33:02 Integrated Voltage Measurement... 10:33:11 Voltage Offset Correction: v_off_ini = -167.359 uV, v_off_fin = -172.829 uV 10:33:11 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -274.234 uVs 10:33:11 Integrated Voltage Measurement... 10:33:20 Voltage Offset Correction: v_off_ini = -142.011 uV, v_off_fin = -142.990 uV 10:33:20 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 275.407 uVs 10:33:20 Integrated Voltage Measurement... 10:33:29 Voltage Offset Correction: v_off_ini = -156.560 uV, v_off_fin = -162.135 uV 10:33:29 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -273.907 uVs 10:33:29 Integrated Voltage Measurement... 10:33:38 Voltage Offset Correction: v_off_ini = -151.569 uV, v_off_fin = -142.035 uV 10:33:38 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 275.937 uVs 10:33:42 I1Y Average and RMS of Measurements VT = -276.859 +- 2.812 uVs I1Y = -55.369 +- 0.566 uTm 10:33:47 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:33:50 Integrated Voltage Measurement... 10:33:59 Voltage Offset Correction: v_off_ini = -148.613 uV, v_off_fin = -185.389 uV 10:33:59 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -52.793 uVs 10:33:59 Integrated Voltage Measurement... 10:34:08 Voltage Offset Correction: v_off_ini = -155.693 uV, v_off_fin = -151.433 uV 10:34:08 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 13.856 uVs 10:34:08 Integrated Voltage Measurement... 10:34:17 Voltage Offset Correction: v_off_ini = -168.195 uV, v_off_fin = -171.375 uV 10:34:17 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -27.354 uVs 10:34:17 Integrated Voltage Measurement... 10:34:26 Voltage Offset Correction: v_off_ini = -160.032 uV, v_off_fin = -161.948 uV 10:34:26 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 13.954 uVs 10:34:26 Integrated Voltage Measurement... 10:34:35 Voltage Offset Correction: v_off_ini = -154.464 uV, v_off_fin = -159.616 uV 10:34:35 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -25.684 uVs 10:34:35 Integrated Voltage Measurement... 10:34:44 Voltage Offset Correction: v_off_ini = -159.025 uV, v_off_fin = -155.653 uV 10:34:44 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 63.203 uVs 10:34:46 I2X Average and RMS of Measurements VT = -32.808 +- 9.719 uVs I2X = 3.018 +- 0.894 uTm^2 10:34:52 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:34:54 Integrated Voltage Measurement... 10:35:03 Voltage Offset Correction: v_off_ini = -146.383 uV, v_off_fin = -151.367 uV 10:35:03 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -295.911 uVs 10:35:03 Integrated Voltage Measurement... 10:35:12 Voltage Offset Correction: v_off_ini = -157.352 uV, v_off_fin = -160.893 uV 10:35:12 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 287.646 uVs 10:35:12 Integrated Voltage Measurement... 10:35:21 Voltage Offset Correction: v_off_ini = -145.816 uV, v_off_fin = -163.764 uV 10:35:21 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -270.088 uVs 10:35:21 Integrated Voltage Measurement... 10:35:30 Voltage Offset Correction: v_off_ini = -147.042 uV, v_off_fin = -140.205 uV 10:35:30 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 305.348 uVs 10:35:30 Integrated Voltage Measurement... 10:35:39 Voltage Offset Correction: v_off_ini = -150.941 uV, v_off_fin = -156.682 uV 10:35:39 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -306.174 uVs 10:35:39 Integrated Voltage Measurement... 10:35:48 Voltage Offset Correction: v_off_ini = -146.749 uV, v_off_fin = -143.895 uV 10:35:48 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 304.111 uVs 10:35:51 I2Y Average and RMS of Measurements VT = -294.880 +- 7.444 uVs I2Y = -27.129 +- 0.685 uTm^2 10:36:01 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:36:06 Integrated Voltage Measurement... 10:36:15 Voltage Offset Correction: v_off_ini = -134.417 uV, v_off_fin = -134.883 uV 10:36:15 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -75.681 uVs 10:36:15 Integrated Voltage Measurement... 10:36:24 Voltage Offset Correction: v_off_ini = -135.200 uV, v_off_fin = -122.733 uV 10:36:24 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 46.154 uVs 10:36:24 Integrated Voltage Measurement... 10:36:33 Voltage Offset Correction: v_off_ini = -137.507 uV, v_off_fin = -152.605 uV 10:36:33 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -59.929 uVs 10:36:33 Integrated Voltage Measurement... 10:36:42 Voltage Offset Correction: v_off_ini = -140.413 uV, v_off_fin = -127.595 uV 10:36:42 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 59.501 uVs 10:36:42 Integrated Voltage Measurement... 10:36:51 Voltage Offset Correction: v_off_ini = -131.841 uV, v_off_fin = -140.440 uV 10:36:51 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -54.196 uVs 10:36:51 Integrated Voltage Measurement... 10:37:00 Voltage Offset Correction: v_off_ini = -148.437 uV, v_off_fin = -150.588 uV 10:37:00 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 73.702 uVs 10:37:04 I1X Average and RMS of Measurements VT = -61.527 +- 1.782 uVs I1X = 12.306 +- 0.356 uTm 10:37:07 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:37:11 Integrated Voltage Measurement... 10:37:20 Voltage Offset Correction: v_off_ini = -131.013 uV, v_off_fin = -137.340 uV 10:37:20 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = -257.269 uVs 10:37:20 Integrated Voltage Measurement... 10:37:29 Voltage Offset Correction: v_off_ini = -149.138 uV, v_off_fin = -123.652 uV 10:37:29 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 276.966 uVs 10:37:29 Integrated Voltage Measurement... 10:37:38 Voltage Offset Correction: v_off_ini = -133.734 uV, v_off_fin = -144.147 uV 10:37:38 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -283.535 uVs 10:37:38 Integrated Voltage Measurement... 10:37:47 Voltage Offset Correction: v_off_ini = -126.094 uV, v_off_fin = -108.055 uV 10:37:47 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 268.599 uVs 10:37:47 Integrated Voltage Measurement... 10:37:56 Voltage Offset Correction: v_off_ini = -118.363 uV, v_off_fin = -140.198 uV 10:37:56 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -258.569 uVs 10:37:56 Integrated Voltage Measurement... 10:38:05 Voltage Offset Correction: v_off_ini = -140.100 uV, v_off_fin = -119.976 uV 10:38:05 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 247.298 uVs 10:38:09 I1Y Average and RMS of Measurements VT = -265.373 +- 9.524 uVs I1Y = -53.071 +- 1.905 uTm 10:38:14 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:38:17 Integrated Voltage Measurement... 10:38:26 Voltage Offset Correction: v_off_ini = -138.458 uV, v_off_fin = -140.189 uV 10:38:26 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -39.557 uVs 10:38:26 Integrated Voltage Measurement... 10:38:35 Voltage Offset Correction: v_off_ini = -161.296 uV, v_off_fin = -150.030 uV 10:38:35 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 7.230 uVs 10:38:35 Integrated Voltage Measurement... 10:38:44 Voltage Offset Correction: v_off_ini = -159.584 uV, v_off_fin = -171.651 uV 10:38:44 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -21.318 uVs 10:38:44 Integrated Voltage Measurement... 10:38:53 Voltage Offset Correction: v_off_ini = -178.524 uV, v_off_fin = -149.460 uV 10:38:53 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 41.770 uVs 10:38:53 Integrated Voltage Measurement... 10:39:02 Voltage Offset Correction: v_off_ini = -153.235 uV, v_off_fin = -167.058 uV 10:39:02 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -27.095 uVs 10:39:02 Integrated Voltage Measurement... 10:39:11 Voltage Offset Correction: v_off_ini = -137.129 uV, v_off_fin = -128.114 uV 10:39:11 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 20.893 uVs 10:39:13 I2X Average and RMS of Measurements VT = -26.311 +- 3.709 uVs I2X = 2.421 +- 0.341 uTm^2 10:39:19 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:39:21 Integrated Voltage Measurement... 10:39:30 Voltage Offset Correction: v_off_ini = -135.243 uV, v_off_fin = -136.234 uV 10:39:30 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -279.844 uVs 10:39:30 Integrated Voltage Measurement... 10:39:39 Voltage Offset Correction: v_off_ini = -128.171 uV, v_off_fin = -146.132 uV 10:39:39 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 276.296 uVs 10:39:39 Integrated Voltage Measurement... 10:39:48 Voltage Offset Correction: v_off_ini = -157.565 uV, v_off_fin = -134.311 uV 10:39:48 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -270.362 uVs 10:39:48 Integrated Voltage Measurement... 10:39:57 Voltage Offset Correction: v_off_ini = -145.942 uV, v_off_fin = -132.876 uV 10:39:57 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 273.637 uVs 10:39:57 Integrated Voltage Measurement... 10:40:06 Voltage Offset Correction: v_off_ini = -132.499 uV, v_off_fin = -136.117 uV 10:40:06 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -278.498 uVs 10:40:06 Integrated Voltage Measurement... 10:40:15 Voltage Offset Correction: v_off_ini = -139.730 uV, v_off_fin = -153.868 uV 10:40:15 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 303.922 uVs 10:40:18 I2Y Average and RMS of Measurements VT = -280.426 +- 8.018 uVs I2Y = -25.799 +- 0.738 uTm^2 10:40:28 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:40:33 Integrated Voltage Measurement... 10:40:42 Voltage Offset Correction: v_off_ini = -159.776 uV, v_off_fin = -154.944 uV 10:40:42 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.4998 mm dXY = -0.4999 mm VT = -51.876 uVs 10:40:42 Integrated Voltage Measurement... 10:40:51 Voltage Offset Correction: v_off_ini = -152.125 uV, v_off_fin = -164.241 uV 10:40:51 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.4999 mm dXY = 0.4999 mm VT = 56.722 uVs 10:40:51 Integrated Voltage Measurement... 10:41:00 Voltage Offset Correction: v_off_ini = -159.326 uV, v_off_fin = -162.231 uV 10:41:00 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -34.306 uVs 10:41:00 Integrated Voltage Measurement... 10:41:09 Voltage Offset Correction: v_off_ini = -142.631 uV, v_off_fin = -140.388 uV 10:41:09 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 43.922 uVs 10:41:09 Integrated Voltage Measurement... 10:41:18 Voltage Offset Correction: v_off_ini = -149.319 uV, v_off_fin = -160.444 uV 10:41:18 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -48.025 uVs 10:41:18 Integrated Voltage Measurement... 10:41:27 Voltage Offset Correction: v_off_ini = -154.559 uV, v_off_fin = -156.865 uV 10:41:27 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 31.170 uVs 10:41:31 I1X Average and RMS of Measurements VT = -44.337 +- 7.047 uVs I1X = 8.868 +- 1.411 uTm 10:41:34 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:41:38 Integrated Voltage Measurement... 10:41:47 Voltage Offset Correction: v_off_ini = -172.742 uV, v_off_fin = -161.946 uV 10:41:47 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -249.681 uVs 10:41:47 Integrated Voltage Measurement... 10:41:56 Voltage Offset Correction: v_off_ini = -147.034 uV, v_off_fin = -144.511 uV 10:41:56 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 261.150 uVs 10:41:56 Integrated Voltage Measurement... 10:42:05 Voltage Offset Correction: v_off_ini = -158.585 uV, v_off_fin = -158.996 uV 10:42:05 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -287.680 uVs 10:42:05 Integrated Voltage Measurement... 10:42:14 Voltage Offset Correction: v_off_ini = -166.631 uV, v_off_fin = -137.774 uV 10:42:14 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 256.002 uVs 10:42:14 Integrated Voltage Measurement... 10:42:23 Voltage Offset Correction: v_off_ini = -151.042 uV, v_off_fin = -160.923 uV 10:42:23 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -246.755 uVs 10:42:23 Integrated Voltage Measurement... 10:42:32 Voltage Offset Correction: v_off_ini = -150.637 uV, v_off_fin = -142.828 uV 10:42:32 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 258.288 uVs 10:42:36 I1Y Average and RMS of Measurements VT = -259.926 +- 8.508 uVs I1Y = -51.983 +- 1.702 uTm 10:42:41 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:42:44 Integrated Voltage Measurement... 10:42:53 Voltage Offset Correction: v_off_ini = -148.550 uV, v_off_fin = -157.614 uV 10:42:53 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -18.959 uVs 10:42:53 Integrated Voltage Measurement... 10:43:02 Voltage Offset Correction: v_off_ini = -151.031 uV, v_off_fin = -134.100 uV 10:43:02 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 21.631 uVs 10:43:02 Integrated Voltage Measurement... 10:43:11 Voltage Offset Correction: v_off_ini = -157.428 uV, v_off_fin = -166.861 uV 10:43:11 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -22.762 uVs 10:43:11 Integrated Voltage Measurement... 10:43:20 Voltage Offset Correction: v_off_ini = -158.602 uV, v_off_fin = -151.961 uV 10:43:20 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 22.785 uVs 10:43:20 Integrated Voltage Measurement... 10:43:29 Voltage Offset Correction: v_off_ini = -143.007 uV, v_off_fin = -161.579 uV 10:43:29 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -28.385 uVs 10:43:29 Integrated Voltage Measurement... 10:43:38 Voltage Offset Correction: v_off_ini = -152.954 uV, v_off_fin = -134.222 uV 10:43:38 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 18.504 uVs 10:43:40 I2X Average and RMS of Measurements VT = -22.171 +- 1.355 uVs I2X = 2.040 +- 0.125 uTm^2 10:43:45 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:43:48 Integrated Voltage Measurement... 10:43:57 Voltage Offset Correction: v_off_ini = -152.446 uV, v_off_fin = -137.080 uV 10:43:57 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -256.531 uVs 10:43:57 Integrated Voltage Measurement... 10:44:06 Voltage Offset Correction: v_off_ini = -137.035 uV, v_off_fin = -137.277 uV 10:44:06 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 274.497 uVs 10:44:06 Integrated Voltage Measurement... 10:44:15 Voltage Offset Correction: v_off_ini = -123.752 uV, v_off_fin = -125.565 uV 10:44:15 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -267.298 uVs 10:44:15 Integrated Voltage Measurement... 10:44:24 Voltage Offset Correction: v_off_ini = -122.953 uV, v_off_fin = -133.418 uV 10:44:24 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 272.029 uVs 10:44:24 Integrated Voltage Measurement... 10:44:33 Voltage Offset Correction: v_off_ini = -125.369 uV, v_off_fin = -130.049 uV 10:44:33 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -281.993 uVs 10:44:33 Integrated Voltage Measurement... 10:44:42 Voltage Offset Correction: v_off_ini = -136.134 uV, v_off_fin = -139.404 uV 10:44:42 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 263.105 uVs 10:44:45 I2Y Average and RMS of Measurements VT = -269.242 +- 2.887 uVs I2Y = -24.770 +- 0.266 uTm^2 10:44:55 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:45:00 Integrated Voltage Measurement... 10:45:09 Voltage Offset Correction: v_off_ini = -120.387 uV, v_off_fin = -123.727 uV 10:45:09 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -47.797 uVs 10:45:09 Integrated Voltage Measurement... 10:45:18 Voltage Offset Correction: v_off_ini = -119.263 uV, v_off_fin = -118.130 uV 10:45:18 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 73.403 uVs 10:45:18 Integrated Voltage Measurement... 10:45:27 Voltage Offset Correction: v_off_ini = -127.526 uV, v_off_fin = -133.791 uV 10:45:27 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -48.758 uVs 10:45:27 Integrated Voltage Measurement... 10:45:36 Voltage Offset Correction: v_off_ini = -119.557 uV, v_off_fin = -111.940 uV 10:45:36 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 47.902 uVs 10:45:36 Integrated Voltage Measurement... 10:45:45 Voltage Offset Correction: v_off_ini = -128.400 uV, v_off_fin = -138.394 uV 10:45:45 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -35.420 uVs 10:45:45 Integrated Voltage Measurement... 10:45:54 Voltage Offset Correction: v_off_ini = -132.069 uV, v_off_fin = -122.498 uV 10:45:54 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 44.308 uVs 10:45:58 I1X Average and RMS of Measurements VT = -49.598 +- 8.513 uVs I1X = 9.920 +- 1.703 uTm 10:46:01 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:46:05 Integrated Voltage Measurement... 10:46:14 Voltage Offset Correction: v_off_ini = -128.319 uV, v_off_fin = -148.128 uV 10:46:14 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -261.008 uVs 10:46:14 Integrated Voltage Measurement... 10:46:23 Voltage Offset Correction: v_off_ini = -132.551 uV, v_off_fin = -112.183 uV 10:46:23 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 272.428 uVs 10:46:23 Integrated Voltage Measurement... 10:46:32 Voltage Offset Correction: v_off_ini = -98.725 uV, v_off_fin = -106.841 uV 10:46:32 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -250.906 uVs 10:46:32 Integrated Voltage Measurement... 10:46:41 Voltage Offset Correction: v_off_ini = -99.666 uV, v_off_fin = -98.841 uV 10:46:41 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 260.550 uVs 10:46:41 Integrated Voltage Measurement... 10:46:50 Voltage Offset Correction: v_off_ini = -105.669 uV, v_off_fin = -103.786 uV 10:46:50 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -255.858 uVs 10:46:50 Integrated Voltage Measurement... 10:46:59 Voltage Offset Correction: v_off_ini = -99.119 uV, v_off_fin = -86.507 uV 10:46:59 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 252.336 uVs 10:47:03 I1Y Average and RMS of Measurements VT = -258.848 +- 5.605 uVs I1Y = -51.769 +- 1.123 uTm 10:47:08 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:47:11 Integrated Voltage Measurement... 10:47:20 Voltage Offset Correction: v_off_ini = -109.315 uV, v_off_fin = -116.617 uV 10:47:20 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -3.375 uVs 10:47:20 Integrated Voltage Measurement... 10:47:29 Voltage Offset Correction: v_off_ini = -103.983 uV, v_off_fin = -92.271 uV 10:47:29 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 9.359 uVs 10:47:29 Integrated Voltage Measurement... 10:47:38 Voltage Offset Correction: v_off_ini = -100.955 uV, v_off_fin = -92.555 uV 10:47:38 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = 5.092 uVs 10:47:38 Integrated Voltage Measurement... 10:47:47 Voltage Offset Correction: v_off_ini = -92.112 uV, v_off_fin = -75.977 uV 10:47:47 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 41.753 uVs 10:47:47 Integrated Voltage Measurement... 10:47:56 Voltage Offset Correction: v_off_ini = -101.180 uV, v_off_fin = -110.698 uV 10:47:56 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -7.659 uVs 10:47:56 Integrated Voltage Measurement... 10:48:05 Voltage Offset Correction: v_off_ini = -82.658 uV, v_off_fin = -70.504 uV 10:48:05 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 9.484 uVs 10:48:07 I2X Average and RMS of Measurements VT = -11.090 +- 5.198 uVs I2X = 1.020 +- 0.478 uTm^2 10:48:13 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:48:15 Integrated Voltage Measurement... 10:48:24 Voltage Offset Correction: v_off_ini = -77.183 uV, v_off_fin = -76.787 uV 10:48:24 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -273.897 uVs 10:48:24 Integrated Voltage Measurement... 10:48:33 Voltage Offset Correction: v_off_ini = -84.159 uV, v_off_fin = -69.859 uV 10:48:33 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 266.025 uVs 10:48:33 Integrated Voltage Measurement... 10:48:42 Voltage Offset Correction: v_off_ini = -89.986 uV, v_off_fin = -82.113 uV 10:48:42 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -260.260 uVs 10:48:42 Integrated Voltage Measurement... 10:48:51 Voltage Offset Correction: v_off_ini = -87.544 uV, v_off_fin = -83.803 uV 10:48:51 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 263.439 uVs 10:48:51 Integrated Voltage Measurement... 10:49:00 Voltage Offset Correction: v_off_ini = -104.253 uV, v_off_fin = -105.248 uV 10:49:00 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -264.586 uVs 10:49:00 Integrated Voltage Measurement... 10:49:09 Voltage Offset Correction: v_off_ini = -74.431 uV, v_off_fin = -92.359 uV 10:49:09 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 266.666 uVs 10:49:12 I2Y Average and RMS of Measurements VT = -265.812 +- 3.314 uVs I2Y = -24.455 +- 0.305 uTm^2 10:49:23 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:49:27 Integrated Voltage Measurement... 10:49:36 Voltage Offset Correction: v_off_ini = -94.002 uV, v_off_fin = -105.977 uV 10:49:36 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4998 mm dXY = -0.4999 mm VT = -34.580 uVs 10:49:36 Integrated Voltage Measurement... 10:49:45 Voltage Offset Correction: v_off_ini = -96.383 uV, v_off_fin = -95.951 uV 10:49:45 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 26.978 uVs 10:49:45 Integrated Voltage Measurement... 10:49:54 Voltage Offset Correction: v_off_ini = -102.824 uV, v_off_fin = -98.209 uV 10:49:54 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -30.127 uVs 10:49:54 Integrated Voltage Measurement... 10:50:03 Voltage Offset Correction: v_off_ini = -87.298 uV, v_off_fin = -85.264 uV 10:50:03 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 25.709 uVs 10:50:03 Integrated Voltage Measurement... 10:50:12 Voltage Offset Correction: v_off_ini = -102.989 uV, v_off_fin = -115.783 uV 10:50:12 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -34.487 uVs 10:50:12 Integrated Voltage Measurement... 10:50:21 Voltage Offset Correction: v_off_ini = -99.394 uV, v_off_fin = -83.229 uV 10:50:21 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 35.825 uVs 10:50:25 I1X Average and RMS of Measurements VT = -31.284 +- 2.976 uVs I1X = 6.257 +- 0.595 uTm 10:50:28 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:50:32 Integrated Voltage Measurement... 10:50:41 Voltage Offset Correction: v_off_ini = -92.570 uV, v_off_fin = -93.736 uV 10:50:41 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = -251.746 uVs 10:50:41 Integrated Voltage Measurement... 10:50:50 Voltage Offset Correction: v_off_ini = -76.080 uV, v_off_fin = -66.655 uV 10:50:50 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 231.820 uVs 10:50:50 Integrated Voltage Measurement... 10:50:59 Voltage Offset Correction: v_off_ini = -73.846 uV, v_off_fin = -84.045 uV 10:50:59 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -251.215 uVs 10:50:59 Integrated Voltage Measurement... 10:51:08 Voltage Offset Correction: v_off_ini = -59.604 uV, v_off_fin = -49.523 uV 10:51:08 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 254.075 uVs 10:51:08 Integrated Voltage Measurement... 10:51:17 Voltage Offset Correction: v_off_ini = -58.068 uV, v_off_fin = -67.040 uV 10:51:17 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -224.173 uVs 10:51:17 Integrated Voltage Measurement... 10:51:26 Voltage Offset Correction: v_off_ini = -58.729 uV, v_off_fin = -51.121 uV 10:51:26 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 241.448 uVs 10:51:30 I1Y Average and RMS of Measurements VT = -242.413 +- 8.110 uVs I1Y = -48.483 +- 1.622 uTm 10:51:35 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:51:38 Integrated Voltage Measurement... 10:51:47 Voltage Offset Correction: v_off_ini = -82.249 uV, v_off_fin = -82.600 uV 10:51:47 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -21.302 uVs 10:51:47 Integrated Voltage Measurement... 10:51:56 Voltage Offset Correction: v_off_ini = -80.568 uV, v_off_fin = -58.060 uV 10:51:56 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 19.109 uVs 10:51:56 Integrated Voltage Measurement... 10:52:05 Voltage Offset Correction: v_off_ini = -68.539 uV, v_off_fin = -89.843 uV 10:52:05 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = 2.066 uVs 10:52:05 Integrated Voltage Measurement... 10:52:14 Voltage Offset Correction: v_off_ini = -78.639 uV, v_off_fin = -71.949 uV 10:52:14 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 33.348 uVs 10:52:14 Integrated Voltage Measurement... 10:52:23 Voltage Offset Correction: v_off_ini = -85.040 uV, v_off_fin = -87.379 uV 10:52:23 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -1.588 uVs 10:52:23 Integrated Voltage Measurement... 10:52:32 Voltage Offset Correction: v_off_ini = -70.337 uV, v_off_fin = -49.455 uV 10:52:32 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 7.275 uVs 10:52:35 I2X Average and RMS of Measurements VT = -13.426 +- 6.627 uVs I2X = 1.235 +- 0.610 uTm^2 10:52:40 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:52:42 Integrated Voltage Measurement... 10:52:51 Voltage Offset Correction: v_off_ini = -64.480 uV, v_off_fin = -72.668 uV 10:52:51 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -260.322 uVs 10:52:51 Integrated Voltage Measurement... 10:53:00 Voltage Offset Correction: v_off_ini = -64.552 uV, v_off_fin = -72.452 uV 10:53:00 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 260.235 uVs 10:53:00 Integrated Voltage Measurement... 10:53:09 Voltage Offset Correction: v_off_ini = -77.342 uV, v_off_fin = -78.444 uV 10:53:09 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -256.194 uVs 10:53:09 Integrated Voltage Measurement... 10:53:18 Voltage Offset Correction: v_off_ini = -66.967 uV, v_off_fin = -61.956 uV 10:53:18 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 259.476 uVs 10:53:18 Integrated Voltage Measurement... 10:53:27 Voltage Offset Correction: v_off_ini = -81.205 uV, v_off_fin = -57.793 uV 10:53:27 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -257.817 uVs 10:53:27 Integrated Voltage Measurement... 10:53:36 Voltage Offset Correction: v_off_ini = -47.896 uV, v_off_fin = -60.548 uV 10:53:37 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 269.994 uVs 10:53:39 I2Y Average and RMS of Measurements VT = -260.673 +- 2.494 uVs I2Y = -23.982 +- 0.229 uTm^2 10:53:50 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:53:54 Integrated Voltage Measurement... 10:54:03 Voltage Offset Correction: v_off_ini = -56.364 uV, v_off_fin = -62.876 uV 10:54:03 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -32.997 uVs 10:54:03 Integrated Voltage Measurement... 10:54:12 Voltage Offset Correction: v_off_ini = -38.600 uV, v_off_fin = -37.386 uV 10:54:12 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = 25.725 uVs 10:54:12 Integrated Voltage Measurement... 10:54:21 Voltage Offset Correction: v_off_ini = -61.215 uV, v_off_fin = -52.865 uV 10:54:21 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -32.997 uVs 10:54:21 Integrated Voltage Measurement... 10:54:30 Voltage Offset Correction: v_off_ini = -34.876 uV, v_off_fin = -41.462 uV 10:54:30 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 30.240 uVs 10:54:30 Integrated Voltage Measurement... 10:54:39 Voltage Offset Correction: v_off_ini = -50.411 uV, v_off_fin = -49.690 uV 10:54:39 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -38.179 uVs 10:54:39 Integrated Voltage Measurement... 10:54:48 Voltage Offset Correction: v_off_ini = -48.791 uV, v_off_fin = -56.108 uV 10:54:48 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 43.165 uVs 10:54:52 I1X Average and RMS of Measurements VT = -33.884 +- 4.887 uVs I1X = 6.777 +- 0.977 uTm 10:54:55 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:54:59 Integrated Voltage Measurement... 10:55:08 Voltage Offset Correction: v_off_ini = -55.852 uV, v_off_fin = -61.501 uV 10:55:08 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -240.368 uVs 10:55:08 Integrated Voltage Measurement... 10:55:17 Voltage Offset Correction: v_off_ini = -43.308 uV, v_off_fin = -31.945 uV 10:55:17 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 230.030 uVs 10:55:17 Integrated Voltage Measurement... 10:55:26 Voltage Offset Correction: v_off_ini = -39.703 uV, v_off_fin = -65.867 uV 10:55:26 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -245.927 uVs 10:55:26 Integrated Voltage Measurement... 10:55:35 Voltage Offset Correction: v_off_ini = -42.384 uV, v_off_fin = -31.340 uV 10:55:35 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 248.284 uVs 10:55:35 Integrated Voltage Measurement... 10:55:44 Voltage Offset Correction: v_off_ini = -48.529 uV, v_off_fin = -59.241 uV 10:55:44 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -237.070 uVs 10:55:44 Integrated Voltage Measurement... 10:55:53 Voltage Offset Correction: v_off_ini = -59.866 uV, v_off_fin = -35.759 uV 10:55:53 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 234.392 uVs 10:55:57 I1Y Average and RMS of Measurements VT = -239.345 +- 5.492 uVs I1Y = -47.869 +- 1.098 uTm 10:56:02 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:56:05 Integrated Voltage Measurement... 10:56:14 Voltage Offset Correction: v_off_ini = -43.742 uV, v_off_fin = -55.695 uV 10:56:14 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -24.088 uVs 10:56:14 Integrated Voltage Measurement... 10:56:23 Voltage Offset Correction: v_off_ini = -43.978 uV, v_off_fin = -39.847 uV 10:56:23 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 7.292 uVs 10:56:23 Integrated Voltage Measurement... 10:56:32 Voltage Offset Correction: v_off_ini = -44.109 uV, v_off_fin = -43.224 uV 10:56:32 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -20.622 uVs 10:56:32 Integrated Voltage Measurement... 10:56:41 Voltage Offset Correction: v_off_ini = -36.033 uV, v_off_fin = -32.155 uV 10:56:41 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 4.419 uVs 10:56:41 Integrated Voltage Measurement... 10:56:50 Voltage Offset Correction: v_off_ini = -41.192 uV, v_off_fin = -54.291 uV 10:56:50 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 0.934 uVs 10:56:50 Integrated Voltage Measurement... 10:56:59 Voltage Offset Correction: v_off_ini = -40.799 uV, v_off_fin = -23.377 uV 10:56:59 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 27.445 uVs 10:57:02 I2X Average and RMS of Measurements VT = -13.822 +- 1.355 uVs I2X = 1.272 +- 0.125 uTm^2 10:57:07 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:57:09 Integrated Voltage Measurement... 10:57:18 Voltage Offset Correction: v_off_ini = -39.404 uV, v_off_fin = -38.784 uV 10:57:18 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -232.123 uVs 10:57:18 Integrated Voltage Measurement... 10:57:27 Voltage Offset Correction: v_off_ini = -24.415 uV, v_off_fin = -35.855 uV 10:57:27 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 241.955 uVs 10:57:27 Integrated Voltage Measurement... 10:57:36 Voltage Offset Correction: v_off_ini = -31.232 uV, v_off_fin = -35.849 uV 10:57:36 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -248.732 uVs 10:57:36 Integrated Voltage Measurement... 10:57:45 Voltage Offset Correction: v_off_ini = -45.132 uV, v_off_fin = -25.316 uV 10:57:45 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 254.324 uVs 10:57:45 Integrated Voltage Measurement... 10:57:54 Voltage Offset Correction: v_off_ini = -33.865 uV, v_off_fin = -37.723 uV 10:57:54 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -257.495 uVs 10:57:54 Integrated Voltage Measurement... 10:58:03 Voltage Offset Correction: v_off_ini = -40.730 uV, v_off_fin = -25.012 uV 10:58:03 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 253.624 uVs 10:58:06 I2Y Average and RMS of Measurements VT = -248.042 +- 7.952 uVs I2Y = -22.820 +- 0.732 uTm^2 10:58:17 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:58:21 Integrated Voltage Measurement... 10:58:30 Voltage Offset Correction: v_off_ini = -21.634 uV, v_off_fin = -38.369 uV 10:58:30 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -52.722 uVs 10:58:30 Integrated Voltage Measurement... 10:58:39 Voltage Offset Correction: v_off_ini = -22.652 uV, v_off_fin = -20.291 uV 10:58:39 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 19.673 uVs 10:58:39 Integrated Voltage Measurement... 10:58:48 Voltage Offset Correction: v_off_ini = -29.315 uV, v_off_fin = -29.213 uV 10:58:48 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -34.197 uVs 10:58:48 Integrated Voltage Measurement... 10:58:57 Voltage Offset Correction: v_off_ini = -25.063 uV, v_off_fin = -21.832 uV 10:58:57 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 36.979 uVs 10:58:57 Integrated Voltage Measurement... 10:59:06 Voltage Offset Correction: v_off_ini = -16.903 uV, v_off_fin = -20.896 uV 10:59:06 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -49.308 uVs 10:59:06 Integrated Voltage Measurement... 10:59:15 Voltage Offset Correction: v_off_ini = -24.455 uV, v_off_fin = -31.401 uV 10:59:15 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 21.805 uVs 10:59:19 I1X Average and RMS of Measurements VT = -35.781 +- 0.295 uVs I1X = 7.156 +- 0.059 uTm 10:59:22 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:59:26 Integrated Voltage Measurement... 10:59:35 Voltage Offset Correction: v_off_ini = -33.653 uV, v_off_fin = -53.038 uV 10:59:35 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -235.271 uVs 10:59:35 Integrated Voltage Measurement... 10:59:44 Voltage Offset Correction: v_off_ini = -37.856 uV, v_off_fin = -37.695 uV 10:59:44 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 234.771 uVs 10:59:44 Integrated Voltage Measurement... 10:59:53 Voltage Offset Correction: v_off_ini = -34.839 uV, v_off_fin = -43.912 uV 10:59:53 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -240.395 uVs 10:59:53 Integrated Voltage Measurement... 11:00:02 Voltage Offset Correction: v_off_ini = -22.868 uV, v_off_fin = -34.911 uV 11:00:02 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 216.544 uVs 11:00:02 Integrated Voltage Measurement... 11:00:11 Voltage Offset Correction: v_off_ini = -30.426 uV, v_off_fin = -42.562 uV 11:00:11 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -238.969 uVs 11:00:11 Integrated Voltage Measurement... 11:00:20 Voltage Offset Correction: v_off_ini = -63.791 uV, v_off_fin = -44.047 uV 11:00:20 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 246.867 uVs 11:00:24 I1Y Average and RMS of Measurements VT = -235.470 +- 5.907 uVs I1Y = -47.092 +- 1.179 uTm 11:00:29 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:00:32 Integrated Voltage Measurement... 11:00:41 Voltage Offset Correction: v_off_ini = -57.041 uV, v_off_fin = -56.415 uV 11:00:41 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -12.204 uVs 11:00:41 Integrated Voltage Measurement... 11:00:50 Voltage Offset Correction: v_off_ini = -61.754 uV, v_off_fin = -41.337 uV 11:00:50 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 2.441 uVs 11:00:50 Integrated Voltage Measurement... 11:00:59 Voltage Offset Correction: v_off_ini = -52.703 uV, v_off_fin = -54.922 uV 11:00:59 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -17.161 uVs 11:00:59 Integrated Voltage Measurement... 11:01:08 Voltage Offset Correction: v_off_ini = -48.593 uV, v_off_fin = -35.655 uV 11:01:08 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 13.358 uVs 11:01:08 Integrated Voltage Measurement... 11:01:17 Voltage Offset Correction: v_off_ini = -43.771 uV, v_off_fin = -45.279 uV 11:01:17 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -25.183 uVs 11:01:17 Integrated Voltage Measurement... 11:01:26 Voltage Offset Correction: v_off_ini = -36.293 uV, v_off_fin = -29.444 uV 11:01:26 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -2.567 uVs 11:01:29 I2X Average and RMS of Measurements VT = -11.297 +- 3.240 uVs I2X = 1.039 +- 0.298 uTm^2 11:01:34 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:01:36 Integrated Voltage Measurement... 11:01:45 Voltage Offset Correction: v_off_ini = -47.436 uV, v_off_fin = -55.036 uV 11:01:45 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -240.257 uVs 11:01:45 Integrated Voltage Measurement... 11:01:54 Voltage Offset Correction: v_off_ini = -59.256 uV, v_off_fin = -57.774 uV 11:01:54 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 258.341 uVs 11:01:54 Integrated Voltage Measurement... 11:02:03 Voltage Offset Correction: v_off_ini = -57.179 uV, v_off_fin = -40.195 uV 11:02:03 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -243.897 uVs 11:02:03 Integrated Voltage Measurement... 11:02:12 Voltage Offset Correction: v_off_ini = -42.987 uV, v_off_fin = -45.504 uV 11:02:12 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 269.675 uVs 11:02:13 Integrated Voltage Measurement... 11:02:21 Voltage Offset Correction: v_off_ini = -43.401 uV, v_off_fin = -45.786 uV 11:02:22 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -272.001 uVs 11:02:22 Integrated Voltage Measurement... 11:02:31 Voltage Offset Correction: v_off_ini = -35.072 uV, v_off_fin = -53.393 uV 11:02:31 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 218.531 uVs 11:02:33 I2Y Average and RMS of Measurements VT = -250.450 +- 4.773 uVs I2Y = -23.041 +- 0.439 uTm^2 11:02:44 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:02:48 Integrated Voltage Measurement... 11:02:57 Voltage Offset Correction: v_off_ini = -45.946 uV, v_off_fin = -51.086 uV 11:02:57 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -33.343 uVs 11:02:57 Integrated Voltage Measurement... 11:03:06 Voltage Offset Correction: v_off_ini = -44.358 uV, v_off_fin = -30.411 uV 11:03:06 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = 31.958 uVs 11:03:06 Integrated Voltage Measurement... 11:03:15 Voltage Offset Correction: v_off_ini = -29.639 uV, v_off_fin = -37.034 uV 11:03:15 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -36.111 uVs 11:03:15 Integrated Voltage Measurement... 11:03:24 Voltage Offset Correction: v_off_ini = -52.272 uV, v_off_fin = -52.990 uV 11:03:24 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 45.261 uVs 11:03:24 Integrated Voltage Measurement... 11:03:33 Voltage Offset Correction: v_off_ini = -46.563 uV, v_off_fin = -47.252 uV 11:03:33 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -58.887 uVs 11:03:33 Integrated Voltage Measurement... 11:03:42 Voltage Offset Correction: v_off_ini = -27.087 uV, v_off_fin = -35.718 uV 11:03:42 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 32.885 uVs 11:03:46 I1X Average and RMS of Measurements VT = -39.741 +- 5.445 uVs I1X = 7.948 +- 1.089 uTm 11:03:49 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:03:53 Integrated Voltage Measurement... 11:04:02 Voltage Offset Correction: v_off_ini = -46.152 uV, v_off_fin = -47.102 uV 11:04:02 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -229.947 uVs 11:04:02 Integrated Voltage Measurement... 11:04:11 Voltage Offset Correction: v_off_ini = -42.654 uV, v_off_fin = -36.034 uV 11:04:11 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 236.577 uVs 11:04:11 Integrated Voltage Measurement... 11:04:20 Voltage Offset Correction: v_off_ini = -35.584 uV, v_off_fin = -53.604 uV 11:04:20 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -239.417 uVs 11:04:20 Integrated Voltage Measurement... 11:04:29 Voltage Offset Correction: v_off_ini = -53.979 uV, v_off_fin = -21.976 uV 11:04:29 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 241.612 uVs 11:04:29 Integrated Voltage Measurement... 11:04:38 Voltage Offset Correction: v_off_ini = -30.740 uV, v_off_fin = -61.278 uV 11:04:38 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -242.705 uVs 11:04:38 Integrated Voltage Measurement... 11:04:47 Voltage Offset Correction: v_off_ini = -34.586 uV, v_off_fin = -21.340 uV 11:04:47 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 224.857 uVs 11:04:51 I1Y Average and RMS of Measurements VT = -235.853 +- 3.303 uVs I1Y = -47.167 +- 0.662 uTm 11:04:56 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:04:59 Integrated Voltage Measurement... 11:05:08 Voltage Offset Correction: v_off_ini = -47.867 uV, v_off_fin = -45.578 uV 11:05:08 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -15.198 uVs 11:05:08 Integrated Voltage Measurement... 11:05:17 Voltage Offset Correction: v_off_ini = -45.761 uV, v_off_fin = -22.413 uV 11:05:17 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 18.322 uVs 11:05:17 Integrated Voltage Measurement... 11:05:26 Voltage Offset Correction: v_off_ini = -27.830 uV, v_off_fin = -46.941 uV 11:05:26 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -38.673 uVs 11:05:26 Integrated Voltage Measurement... 11:05:35 Voltage Offset Correction: v_off_ini = -31.940 uV, v_off_fin = -27.934 uV 11:05:35 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = -15.121 uVs 11:05:35 Integrated Voltage Measurement... 11:05:44 Voltage Offset Correction: v_off_ini = -27.544 uV, v_off_fin = -39.086 uV 11:05:44 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -7.632 uVs 11:05:44 Integrated Voltage Measurement... 11:05:53 Voltage Offset Correction: v_off_ini = -26.186 uV, v_off_fin = -10.072 uV 11:05:53 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 21.509 uVs 11:05:56 I2X Average and RMS of Measurements VT = -14.369 +- 2.040 uVs I2X = 1.322 +- 0.188 uTm^2 11:06:01 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:06:03 Integrated Voltage Measurement... 11:06:12 Voltage Offset Correction: v_off_ini = -17.909 uV, v_off_fin = -20.073 uV 11:06:12 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -241.301 uVs 11:06:12 Integrated Voltage Measurement... 11:06:21 Voltage Offset Correction: v_off_ini = -12.001 uV, v_off_fin = -20.216 uV 11:06:21 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 244.921 uVs 11:06:22 Integrated Voltage Measurement... 11:06:30 Voltage Offset Correction: v_off_ini = -2.447 uV, v_off_fin = -9.614 uV 11:06:31 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -240.083 uVs 11:06:31 Integrated Voltage Measurement... 11:06:40 Voltage Offset Correction: v_off_ini = 2.760 uV, v_off_fin = -7.094 uV 11:06:40 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 222.169 uVs 11:06:40 Integrated Voltage Measurement... 11:06:49 Voltage Offset Correction: v_off_ini = 4.508 uV, v_off_fin = -6.592 uV 11:06:49 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -242.495 uVs 11:06:49 Integrated Voltage Measurement... 11:06:58 Voltage Offset Correction: v_off_ini = -14.342 uV, v_off_fin = -18.575 uV 11:06:58 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 248.632 uVs 11:07:00 I2Y Average and RMS of Measurements VT = -239.934 +- 6.308 uVs I2Y = -22.074 +- 0.580 uTm^2 11:07:12 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:07:16 Integrated Voltage Measurement... 11:07:25 Voltage Offset Correction: v_off_ini = 0.302 uV, v_off_fin = -5.147 uV 11:07:25 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.4998 mm dXY = -0.4999 mm VT = -37.949 uVs 11:07:25 Integrated Voltage Measurement... 11:07:34 Voltage Offset Correction: v_off_ini = -5.438 uV, v_off_fin = 6.940 uV 11:07:34 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 5.751 uVs 11:07:34 Integrated Voltage Measurement... 11:07:43 Voltage Offset Correction: v_off_ini = -0.121 uV, v_off_fin = 3.673 uV 11:07:43 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -27.759 uVs 11:07:43 Integrated Voltage Measurement... 11:07:52 Voltage Offset Correction: v_off_ini = -5.943 uV, v_off_fin = 19.509 uV 11:07:52 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 17.202 uVs 11:07:52 Integrated Voltage Measurement... 11:08:01 Voltage Offset Correction: v_off_ini = 7.626 uV, v_off_fin = -2.432 uV 11:08:01 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -30.985 uVs 11:08:01 Integrated Voltage Measurement... 11:08:10 Voltage Offset Correction: v_off_ini = 0.385 uV, v_off_fin = 2.558 uV 11:08:10 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 30.513 uVs 11:08:15 I1X Average and RMS of Measurements VT = -25.026 +- 4.054 uVs I1X = 5.006 +- 0.811 uTm 11:08:17 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:08:21 Integrated Voltage Measurement... 11:08:30 Voltage Offset Correction: v_off_ini = -4.933 uV, v_off_fin = -14.714 uV 11:08:30 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.4999 mm dXY = 0.4999 mm VT = -218.076 uVs 11:08:30 Integrated Voltage Measurement... 11:08:39 Voltage Offset Correction: v_off_ini = -7.582 uV, v_off_fin = 12.495 uV 11:08:39 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 212.254 uVs 11:08:39 Integrated Voltage Measurement... 11:08:48 Voltage Offset Correction: v_off_ini = 5.541 uV, v_off_fin = -11.924 uV 11:08:48 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -230.014 uVs 11:08:48 Integrated Voltage Measurement... 11:08:57 Voltage Offset Correction: v_off_ini = 19.362 uV, v_off_fin = 28.131 uV 11:08:57 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 219.866 uVs 11:08:57 Integrated Voltage Measurement... 11:09:06 Voltage Offset Correction: v_off_ini = 10.902 uV, v_off_fin = -13.429 uV 11:09:06 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -245.451 uVs 11:09:06 Integrated Voltage Measurement... 11:09:15 Voltage Offset Correction: v_off_ini = -5.603 uV, v_off_fin = 14.168 uV 11:09:15 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 220.546 uVs 11:09:20 I1Y Average and RMS of Measurements VT = -224.368 +- 7.292 uVs I1Y = -44.874 +- 1.457 uTm 11:09:25 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:09:27 Integrated Voltage Measurement... 11:09:36 Voltage Offset Correction: v_off_ini = 3.392 uV, v_off_fin = -4.317 uV 11:09:36 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -8.124 uVs 11:09:36 Integrated Voltage Measurement... 11:09:45 Voltage Offset Correction: v_off_ini = 21.900 uV, v_off_fin = 15.018 uV 11:09:45 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 39.280 uVs 11:09:45 Integrated Voltage Measurement... 11:09:54 Voltage Offset Correction: v_off_ini = -11.481 uV, v_off_fin = -15.933 uV 11:09:54 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 4.100 uVs 11:09:54 Integrated Voltage Measurement... 11:10:03 Voltage Offset Correction: v_off_ini = 1.039 uV, v_off_fin = -3.493 uV 11:10:03 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 28.533 uVs 11:10:04 Integrated Voltage Measurement... 11:10:12 Voltage Offset Correction: v_off_ini = 1.913 uV, v_off_fin = -9.605 uV 11:10:13 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -1.862 uVs 11:10:13 Integrated Voltage Measurement... 11:10:21 Voltage Offset Correction: v_off_ini = -0.299 uV, v_off_fin = -4.455 uV 11:10:22 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 2.721 uVs 11:10:24 I2X Average and RMS of Measurements VT = -12.737 +- 8.749 uVs I2X = 1.172 +- 0.805 uTm^2 11:10:29 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:10:32 Integrated Voltage Measurement... 11:10:41 Voltage Offset Correction: v_off_ini = -9.642 uV, v_off_fin = -26.026 uV 11:10:41 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -252.803 uVs 11:10:41 Integrated Voltage Measurement... 11:10:50 Voltage Offset Correction: v_off_ini = -10.262 uV, v_off_fin = -10.532 uV 11:10:50 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 230.649 uVs 11:10:50 Integrated Voltage Measurement... 11:10:59 Voltage Offset Correction: v_off_ini = -19.022 uV, v_off_fin = -5.782 uV 11:10:59 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -237.900 uVs 11:10:59 Integrated Voltage Measurement... 11:11:08 Voltage Offset Correction: v_off_ini = -15.453 uV, v_off_fin = -6.189 uV 11:11:08 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 215.926 uVs 11:11:08 Integrated Voltage Measurement... 11:11:17 Voltage Offset Correction: v_off_ini = -17.711 uV, v_off_fin = -12.992 uV 11:11:17 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -239.032 uVs 11:11:17 Integrated Voltage Measurement... 11:11:26 Voltage Offset Correction: v_off_ini = 0.353 uV, v_off_fin = 2.350 uV 11:11:26 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 216.709 uVs 11:11:28 I2Y Average and RMS of Measurements VT = -232.170 +- 6.769 uVs I2Y = -21.360 +- 0.623 uTm^2 11:11:40 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:11:44 Integrated Voltage Measurement... 11:11:53 Voltage Offset Correction: v_off_ini = -8.527 uV, v_off_fin = -8.849 uV 11:11:53 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -33.644 uVs 11:11:53 Integrated Voltage Measurement... 11:12:02 Voltage Offset Correction: v_off_ini = 2.425 uV, v_off_fin = -2.407 uV 11:12:02 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 29.798 uVs 11:12:02 Integrated Voltage Measurement... 11:12:11 Voltage Offset Correction: v_off_ini = -14.578 uV, v_off_fin = -24.970 uV 11:12:11 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -8.259 uVs 11:12:11 Integrated Voltage Measurement... 11:12:20 Voltage Offset Correction: v_off_ini = -13.123 uV, v_off_fin = -0.168 uV 11:12:20 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 39.345 uVs 11:12:20 Integrated Voltage Measurement... 11:12:29 Voltage Offset Correction: v_off_ini = -14.612 uV, v_off_fin = -4.939 uV 11:12:29 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -21.804 uVs 11:12:29 Integrated Voltage Measurement... 11:12:38 Voltage Offset Correction: v_off_ini = -9.222 uV, v_off_fin = 1.309 uV 11:12:38 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 28.883 uVs 11:12:43 I1X Average and RMS of Measurements VT = -26.956 +- 3.428 uVs I1X = 5.391 +- 0.686 uTm 11:12:45 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:12:49 Integrated Voltage Measurement... 11:12:58 Voltage Offset Correction: v_off_ini = -0.605 uV, v_off_fin = -7.337 uV 11:12:58 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = -216.468 uVs 11:12:58 Integrated Voltage Measurement... 11:13:07 Voltage Offset Correction: v_off_ini = -1.649 uV, v_off_fin = 31.020 uV 11:13:07 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 206.023 uVs 11:13:07 Integrated Voltage Measurement... 11:13:16 Voltage Offset Correction: v_off_ini = 25.107 uV, v_off_fin = 5.912 uV 11:13:16 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -217.039 uVs 11:13:16 Integrated Voltage Measurement... 11:13:25 Voltage Offset Correction: v_off_ini = 29.687 uV, v_off_fin = 16.065 uV 11:13:25 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 212.321 uVs 11:13:25 Integrated Voltage Measurement... 11:13:34 Voltage Offset Correction: v_off_ini = 11.800 uV, v_off_fin = 8.427 uV 11:13:34 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -214.842 uVs 11:13:34 Integrated Voltage Measurement... 11:13:43 Voltage Offset Correction: v_off_ini = 22.941 uV, v_off_fin = 22.405 uV 11:13:43 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 210.812 uVs 11:13:48 I1Y Average and RMS of Measurements VT = -212.918 +- 1.404 uVs I1Y = -42.583 +- 0.279 uTm 11:13:53 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:13:55 Integrated Voltage Measurement... 11:14:04 Voltage Offset Correction: v_off_ini = 14.381 uV, v_off_fin = 3.078 uV 11:14:04 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 15.203 uVs 11:14:04 Integrated Voltage Measurement... 11:14:13 Voltage Offset Correction: v_off_ini = 19.975 uV, v_off_fin = 29.477 uV 11:14:13 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -20.312 uVs 11:14:14 Integrated Voltage Measurement... 11:14:23 Voltage Offset Correction: v_off_ini = 22.175 uV, v_off_fin = 12.475 uV 11:14:23 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -10.975 uVs 11:14:23 Integrated Voltage Measurement... 11:14:32 Voltage Offset Correction: v_off_ini = 45.173 uV, v_off_fin = 55.725 uV 11:14:32 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = -6.184 uVs 11:14:32 Integrated Voltage Measurement... 11:14:41 Voltage Offset Correction: v_off_ini = 38.612 uV, v_off_fin = 20.007 uV 11:14:41 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -18.953 uVs 11:14:41 Integrated Voltage Measurement... 11:14:50 Voltage Offset Correction: v_off_ini = 32.555 uV, v_off_fin = 37.626 uV 11:14:50 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 9.881 uVs 11:14:52 I2X Average and RMS of Measurements VT = 0.315 +- 13.274 uVs I2X = -0.029 +- 1.221 uTm^2 11:14:57 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:15:00 Integrated Voltage Measurement... 11:15:09 Voltage Offset Correction: v_off_ini = 34.820 uV, v_off_fin = 35.122 uV 11:15:09 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -231.037 uVs 11:15:09 Integrated Voltage Measurement... 11:15:18 Voltage Offset Correction: v_off_ini = 48.467 uV, v_off_fin = 56.427 uV 11:15:18 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 226.296 uVs 11:15:18 Integrated Voltage Measurement... 11:15:27 Voltage Offset Correction: v_off_ini = 35.434 uV, v_off_fin = 47.249 uV 11:15:27 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -194.097 uVs 11:15:27 Integrated Voltage Measurement... 11:15:36 Voltage Offset Correction: v_off_ini = 41.906 uV, v_off_fin = 44.957 uV 11:15:36 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 228.975 uVs 11:15:36 Integrated Voltage Measurement... 11:15:45 Voltage Offset Correction: v_off_ini = 60.838 uV, v_off_fin = 64.178 uV 11:15:45 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -245.223 uVs 11:15:45 Integrated Voltage Measurement... 11:15:54 Voltage Offset Correction: v_off_ini = 55.299 uV, v_off_fin = 58.595 uV 11:15:54 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 205.558 uVs 11:15:56 I2Y Average and RMS of Measurements VT = -221.864 +- 7.425 uVs I2Y = -20.412 +- 0.683 uTm^2 11:16:10 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:16:15 Integrated Voltage Measurement... 11:16:24 Voltage Offset Correction: v_off_ini = 41.154 uV, v_off_fin = 51.629 uV 11:16:24 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -36.511 uVs 11:16:24 Integrated Voltage Measurement... 11:16:33 Voltage Offset Correction: v_off_ini = 55.998 uV, v_off_fin = 57.694 uV 11:16:33 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 16.688 uVs 11:16:33 Integrated Voltage Measurement... 11:16:42 Voltage Offset Correction: v_off_ini = 62.775 uV, v_off_fin = 53.503 uV 11:16:42 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -56.254 uVs 11:16:42 Integrated Voltage Measurement... 11:16:51 Voltage Offset Correction: v_off_ini = 57.469 uV, v_off_fin = 45.922 uV 11:16:51 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 18.975 uVs 11:16:51 Integrated Voltage Measurement... 11:17:00 Voltage Offset Correction: v_off_ini = 36.081 uV, v_off_fin = 44.647 uV 11:17:00 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -25.307 uVs 11:17:00 Integrated Voltage Measurement... 11:17:09 Voltage Offset Correction: v_off_ini = 38.375 uV, v_off_fin = 39.560 uV 11:17:09 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 53.248 uVs 11:17:13 I1X Average and RMS of Measurements VT = -34.497 +- 5.625 uVs I1X = 6.899 +- 1.125 uTm 11:17:16 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:17:20 Integrated Voltage Measurement... 11:17:29 Voltage Offset Correction: v_off_ini = 28.955 uV, v_off_fin = 20.350 uV 11:17:29 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = -187.763 uVs 11:17:29 Integrated Voltage Measurement... 11:17:38 Voltage Offset Correction: v_off_ini = 14.921 uV, v_off_fin = 49.753 uV 11:17:38 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 194.255 uVs 11:17:38 Integrated Voltage Measurement... 11:17:47 Voltage Offset Correction: v_off_ini = 35.958 uV, v_off_fin = 16.263 uV 11:17:47 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -197.064 uVs 11:17:47 Integrated Voltage Measurement... 11:17:56 Voltage Offset Correction: v_off_ini = 29.315 uV, v_off_fin = 43.154 uV 11:17:56 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 194.778 uVs 11:17:56 Integrated Voltage Measurement... 11:18:05 Voltage Offset Correction: v_off_ini = 45.268 uV, v_off_fin = 32.480 uV 11:18:05 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -202.818 uVs 11:18:05 Integrated Voltage Measurement... 11:18:14 Voltage Offset Correction: v_off_ini = 53.065 uV, v_off_fin = 59.356 uV 11:18:14 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 202.146 uVs 11:18:18 I1Y Average and RMS of Measurements VT = -196.471 +- 4.700 uVs I1Y = -39.294 +- 0.939 uTm 11:18:23 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:18:26 Integrated Voltage Measurement... 11:18:35 Voltage Offset Correction: v_off_ini = 39.455 uV, v_off_fin = 27.578 uV 11:18:35 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -3.692 uVs 11:18:35 Integrated Voltage Measurement... 11:18:44 Voltage Offset Correction: v_off_ini = 36.420 uV, v_off_fin = 50.535 uV 11:18:44 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 9.089 uVs 11:18:44 Integrated Voltage Measurement... 11:18:53 Voltage Offset Correction: v_off_ini = 39.884 uV, v_off_fin = 22.388 uV 11:18:53 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -7.370 uVs 11:18:53 Integrated Voltage Measurement... 11:19:02 Voltage Offset Correction: v_off_ini = 35.337 uV, v_off_fin = 43.928 uV 11:19:02 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 8.285 uVs 11:19:02 Integrated Voltage Measurement... 11:19:11 Voltage Offset Correction: v_off_ini = 33.895 uV, v_off_fin = 28.280 uV 11:19:11 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -10.978 uVs 11:19:11 Integrated Voltage Measurement... 11:19:20 Voltage Offset Correction: v_off_ini = 46.333 uV, v_off_fin = 53.179 uV 11:19:20 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 12.909 uVs 11:19:22 I2X Average and RMS of Measurements VT = -8.720 +- 2.353 uVs I2X = 0.802 +- 0.217 uTm^2 11:19:28 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:19:30 Integrated Voltage Measurement... 11:19:39 Voltage Offset Correction: v_off_ini = 40.079 uV, v_off_fin = 48.914 uV 11:19:39 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -190.887 uVs 11:19:39 Integrated Voltage Measurement... 11:19:48 Voltage Offset Correction: v_off_ini = 44.503 uV, v_off_fin = 55.694 uV 11:19:48 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 198.436 uVs 11:19:48 Integrated Voltage Measurement... 11:19:57 Voltage Offset Correction: v_off_ini = 52.698 uV, v_off_fin = 49.561 uV 11:19:57 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -200.409 uVs 11:19:57 Integrated Voltage Measurement... 11:20:06 Voltage Offset Correction: v_off_ini = 59.756 uV, v_off_fin = 57.794 uV 11:20:06 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 201.669 uVs 11:20:06 Integrated Voltage Measurement... 11:20:15 Voltage Offset Correction: v_off_ini = 57.394 uV, v_off_fin = 60.326 uV 11:20:15 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -205.191 uVs 11:20:15 Integrated Voltage Measurement... 11:20:24 Voltage Offset Correction: v_off_ini = 58.306 uV, v_off_fin = 53.698 uV 11:20:24 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 201.239 uVs 11:20:27 I2Y Average and RMS of Measurements VT = -199.639 +- 3.630 uVs I2Y = -18.367 +- 0.334 uTm^2 11:20:41 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:20:45 Integrated Voltage Measurement... 11:20:54 Voltage Offset Correction: v_off_ini = 63.950 uV, v_off_fin = 69.312 uV 11:20:54 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -48.893 uVs 11:20:54 Integrated Voltage Measurement... 11:21:03 Voltage Offset Correction: v_off_ini = 54.810 uV, v_off_fin = 72.750 uV 11:21:03 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 26.689 uVs 11:21:03 Integrated Voltage Measurement... 11:21:12 Voltage Offset Correction: v_off_ini = 67.033 uV, v_off_fin = 57.149 uV 11:21:12 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -22.839 uVs 11:21:12 Integrated Voltage Measurement... 11:21:21 Voltage Offset Correction: v_off_ini = 59.451 uV, v_off_fin = 54.679 uV 11:21:21 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 16.264 uVs 11:21:21 Integrated Voltage Measurement... 11:21:30 Voltage Offset Correction: v_off_ini = 59.626 uV, v_off_fin = 59.075 uV 11:21:30 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -28.972 uVs 11:21:30 Integrated Voltage Measurement... 11:21:39 Voltage Offset Correction: v_off_ini = 74.601 uV, v_off_fin = 81.551 uV 11:21:39 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 21.067 uVs 11:21:43 I1X Average and RMS of Measurements VT = -27.454 +- 7.643 uVs I1X = 5.491 +- 1.529 uTm 11:21:46 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:21:50 Integrated Voltage Measurement... 11:21:59 Voltage Offset Correction: v_off_ini = 72.753 uV, v_off_fin = 53.049 uV 11:21:59 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = -182.053 uVs 11:21:59 Integrated Voltage Measurement... 11:22:08 Voltage Offset Correction: v_off_ini = 58.602 uV, v_off_fin = 72.667 uV 11:22:08 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 195.648 uVs 11:22:08 Integrated Voltage Measurement... 11:22:17 Voltage Offset Correction: v_off_ini = 75.616 uV, v_off_fin = 43.692 uV 11:22:17 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -184.096 uVs 11:22:17 Integrated Voltage Measurement... 11:22:26 Voltage Offset Correction: v_off_ini = 62.716 uV, v_off_fin = 90.462 uV 11:22:26 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 191.999 uVs 11:22:26 Integrated Voltage Measurement... 11:22:35 Voltage Offset Correction: v_off_ini = 73.199 uV, v_off_fin = 65.861 uV 11:22:35 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -188.481 uVs 11:22:35 Integrated Voltage Measurement... 11:22:44 Voltage Offset Correction: v_off_ini = 69.967 uV, v_off_fin = 75.553 uV 11:22:44 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 203.881 uVs 11:22:48 I1Y Average and RMS of Measurements VT = -191.026 +- 3.660 uVs I1Y = -38.206 +- 0.732 uTm 11:22:54 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:22:56 Integrated Voltage Measurement... 11:23:05 Voltage Offset Correction: v_off_ini = 61.634 uV, v_off_fin = 60.757 uV 11:23:05 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -2.751 uVs 11:23:05 Integrated Voltage Measurement... 11:23:14 Voltage Offset Correction: v_off_ini = 69.839 uV, v_off_fin = 67.583 uV 11:23:14 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 23.419 uVs 11:23:14 Integrated Voltage Measurement... 11:23:23 Voltage Offset Correction: v_off_ini = 65.608 uV, v_off_fin = 60.885 uV 11:23:23 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -19.095 uVs 11:23:23 Integrated Voltage Measurement... 11:23:32 Voltage Offset Correction: v_off_ini = 66.991 uV, v_off_fin = 84.358 uV 11:23:32 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 2.801 uVs 11:23:32 Integrated Voltage Measurement... 11:23:41 Voltage Offset Correction: v_off_ini = 80.459 uV, v_off_fin = 71.160 uV 11:23:41 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -9.863 uVs 11:23:41 Integrated Voltage Measurement... 11:23:50 Voltage Offset Correction: v_off_ini = 69.006 uV, v_off_fin = 86.482 uV 11:23:50 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 19.677 uVs 11:23:53 I2X Average and RMS of Measurements VT = -12.934 +- 1.564 uVs I2X = 1.190 +- 0.144 uTm^2 11:23:58 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:24:00 Integrated Voltage Measurement... 11:24:09 Voltage Offset Correction: v_off_ini = 67.983 uV, v_off_fin = 76.229 uV 11:24:10 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -183.119 uVs 11:24:10 Integrated Voltage Measurement... 11:24:19 Voltage Offset Correction: v_off_ini = 77.227 uV, v_off_fin = 81.867 uV 11:24:19 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 192.069 uVs 11:24:19 Integrated Voltage Measurement... 11:24:28 Voltage Offset Correction: v_off_ini = 81.068 uV, v_off_fin = 77.752 uV 11:24:28 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -181.148 uVs 11:24:28 Integrated Voltage Measurement... 11:24:37 Voltage Offset Correction: v_off_ini = 82.313 uV, v_off_fin = 83.472 uV 11:24:37 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 188.646 uVs 11:24:37 Integrated Voltage Measurement... 11:24:46 Voltage Offset Correction: v_off_ini = 85.684 uV, v_off_fin = 84.259 uV 11:24:46 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -193.728 uVs 11:24:46 Integrated Voltage Measurement... 11:24:55 Voltage Offset Correction: v_off_ini = 78.286 uV, v_off_fin = 82.970 uV 11:24:55 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 210.136 uVs 11:24:57 I2Y Average and RMS of Measurements VT = -191.474 +- 7.476 uVs I2Y = -17.616 +- 0.688 uTm^2 11:25:11 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:25:15 Integrated Voltage Measurement... 11:25:24 Voltage Offset Correction: v_off_ini = 71.197 uV, v_off_fin = 84.424 uV 11:25:24 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -50.682 uVs 11:25:24 Integrated Voltage Measurement... 11:25:33 Voltage Offset Correction: v_off_ini = 75.176 uV, v_off_fin = 89.757 uV 11:25:33 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 39.419 uVs 11:25:34 Integrated Voltage Measurement... 11:25:42 Voltage Offset Correction: v_off_ini = 66.022 uV, v_off_fin = 72.722 uV 11:25:42 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -25.850 uVs 11:25:43 Integrated Voltage Measurement... 11:25:51 Voltage Offset Correction: v_off_ini = 91.035 uV, v_off_fin = 101.296 uV 11:25:52 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 29.539 uVs 11:25:52 Integrated Voltage Measurement... 11:26:01 Voltage Offset Correction: v_off_ini = 95.808 uV, v_off_fin = 102.078 uV 11:26:01 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -51.717 uVs 11:26:01 Integrated Voltage Measurement... 11:26:10 Voltage Offset Correction: v_off_ini = 100.649 uV, v_off_fin = 112.712 uV 11:26:10 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 45.329 uVs 11:26:14 I1X Average and RMS of Measurements VT = -40.423 +- 9.111 uVs I1X = 8.085 +- 1.822 uTm 11:26:16 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:26:21 Integrated Voltage Measurement... 11:26:30 Voltage Offset Correction: v_off_ini = 86.612 uV, v_off_fin = 66.101 uV 11:26:30 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -168.830 uVs 11:26:30 Integrated Voltage Measurement... 11:26:38 Voltage Offset Correction: v_off_ini = 77.908 uV, v_off_fin = 89.759 uV 11:26:39 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 185.603 uVs 11:26:39 Integrated Voltage Measurement... 11:26:48 Voltage Offset Correction: v_off_ini = 88.973 uV, v_off_fin = 78.917 uV 11:26:48 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -174.966 uVs 11:26:48 Integrated Voltage Measurement... 11:26:57 Voltage Offset Correction: v_off_ini = 90.872 uV, v_off_fin = 110.468 uV 11:26:57 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 196.470 uVs 11:26:57 Integrated Voltage Measurement... 11:27:06 Voltage Offset Correction: v_off_ini = 81.610 uV, v_off_fin = 79.510 uV 11:27:06 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -183.514 uVs 11:27:06 Integrated Voltage Measurement... 11:27:15 Voltage Offset Correction: v_off_ini = 89.970 uV, v_off_fin = 112.383 uV 11:27:15 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 198.516 uVs 11:27:19 I1Y Average and RMS of Measurements VT = -184.650 +- 5.684 uVs I1Y = -36.931 +- 1.137 uTm 11:27:24 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:27:26 Integrated Voltage Measurement... 11:27:35 Voltage Offset Correction: v_off_ini = 75.585 uV, v_off_fin = 80.698 uV 11:27:36 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -27.167 uVs 11:27:36 Integrated Voltage Measurement... 11:27:45 Voltage Offset Correction: v_off_ini = 87.705 uV, v_off_fin = 87.961 uV 11:27:45 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 2.238 uVs 11:27:45 Integrated Voltage Measurement... 11:27:54 Voltage Offset Correction: v_off_ini = 82.741 uV, v_off_fin = 76.877 uV 11:27:54 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -13.399 uVs 11:27:54 Integrated Voltage Measurement... 11:28:03 Voltage Offset Correction: v_off_ini = 85.062 uV, v_off_fin = 97.460 uV 11:28:03 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 7.922 uVs 11:28:03 Integrated Voltage Measurement... 11:28:12 Voltage Offset Correction: v_off_ini = 79.536 uV, v_off_fin = 75.815 uV 11:28:12 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -10.547 uVs 11:28:12 Integrated Voltage Measurement... 11:28:21 Voltage Offset Correction: v_off_ini = 84.396 uV, v_off_fin = 84.089 uV 11:28:21 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 35.033 uVs 11:28:23 I2X Average and RMS of Measurements VT = -16.051 +- 5.043 uVs I2X = 1.477 +- 0.464 uTm^2 11:28:28 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:28:31 Integrated Voltage Measurement... 11:28:40 Voltage Offset Correction: v_off_ini = 95.424 uV, v_off_fin = 92.971 uV 11:28:40 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = -209.343 uVs 11:28:40 Integrated Voltage Measurement... 11:28:49 Voltage Offset Correction: v_off_ini = 90.268 uV, v_off_fin = 79.795 uV 11:28:49 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 180.066 uVs 11:28:49 Integrated Voltage Measurement... 11:28:58 Voltage Offset Correction: v_off_ini = 89.897 uV, v_off_fin = 96.180 uV 11:28:58 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -166.002 uVs 11:28:58 Integrated Voltage Measurement... 11:29:07 Voltage Offset Correction: v_off_ini = 92.992 uV, v_off_fin = 91.025 uV 11:29:07 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 167.854 uVs 11:29:07 Integrated Voltage Measurement... 11:29:16 Voltage Offset Correction: v_off_ini = 90.990 uV, v_off_fin = 80.213 uV 11:29:16 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -180.280 uVs 11:29:16 Integrated Voltage Measurement... 11:29:25 Voltage Offset Correction: v_off_ini = 83.687 uV, v_off_fin = 86.964 uV 11:29:25 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 173.761 uVs 11:29:27 I2Y Average and RMS of Measurements VT = -179.551 +- 11.480 uVs I2Y = -16.519 +- 1.057 uTm^2 11:29:42 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:29:46 Integrated Voltage Measurement... 11:29:55 Voltage Offset Correction: v_off_ini = 99.099 uV, v_off_fin = 89.106 uV 11:29:55 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -65.722 uVs 11:29:55 Integrated Voltage Measurement... 11:30:04 Voltage Offset Correction: v_off_ini = 89.321 uV, v_off_fin = 105.823 uV 11:30:04 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 37.042 uVs 11:30:04 Integrated Voltage Measurement... 11:30:13 Voltage Offset Correction: v_off_ini = 92.771 uV, v_off_fin = 85.626 uV 11:30:13 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -57.230 uVs 11:30:13 Integrated Voltage Measurement... 11:30:22 Voltage Offset Correction: v_off_ini = 87.296 uV, v_off_fin = 100.594 uV 11:30:22 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 47.548 uVs 11:30:22 Integrated Voltage Measurement... 11:30:31 Voltage Offset Correction: v_off_ini = 83.019 uV, v_off_fin = 84.405 uV 11:30:31 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -37.641 uVs 11:30:31 Integrated Voltage Measurement... 11:30:40 Voltage Offset Correction: v_off_ini = 78.372 uV, v_off_fin = 86.669 uV 11:30:40 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 56.683 uVs 11:30:44 I1X Average and RMS of Measurements VT = -50.311 +- 2.264 uVs I1X = 10.062 +- 0.453 uTm 11:30:47 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:30:51 Integrated Voltage Measurement... 11:31:00 Voltage Offset Correction: v_off_ini = 87.864 uV, v_off_fin = 75.304 uV 11:31:00 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -195.275 uVs 11:31:00 Integrated Voltage Measurement... 11:31:09 Voltage Offset Correction: v_off_ini = 106.820 uV, v_off_fin = 111.793 uV 11:31:09 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 196.070 uVs 11:31:09 Integrated Voltage Measurement... 11:31:18 Voltage Offset Correction: v_off_ini = 97.503 uV, v_off_fin = 82.435 uV 11:31:18 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -180.345 uVs 11:31:18 Integrated Voltage Measurement... 11:31:27 Voltage Offset Correction: v_off_ini = 88.828 uV, v_off_fin = 117.739 uV 11:31:27 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 162.170 uVs 11:31:27 Integrated Voltage Measurement... 11:31:36 Voltage Offset Correction: v_off_ini = 95.861 uV, v_off_fin = 86.334 uV 11:31:36 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -180.279 uVs 11:31:36 Integrated Voltage Measurement... 11:31:45 Voltage Offset Correction: v_off_ini = 86.261 uV, v_off_fin = 109.468 uV 11:31:45 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 185.662 uVs 11:31:49 I1Y Average and RMS of Measurements VT = -183.300 +- 9.970 uVs I1Y = -36.660 +- 1.994 uTm 11:31:54 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:31:57 Integrated Voltage Measurement... 11:32:06 Voltage Offset Correction: v_off_ini = 82.710 uV, v_off_fin = 90.218 uV 11:32:06 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -12.417 uVs 11:32:06 Integrated Voltage Measurement... 11:32:15 Voltage Offset Correction: v_off_ini = 101.974 uV, v_off_fin = 103.913 uV 11:32:15 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 37.105 uVs 11:32:15 Integrated Voltage Measurement... 11:32:24 Voltage Offset Correction: v_off_ini = 106.794 uV, v_off_fin = 94.760 uV 11:32:24 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -35.419 uVs 11:32:24 Integrated Voltage Measurement... 11:32:33 Voltage Offset Correction: v_off_ini = 102.031 uV, v_off_fin = 121.954 uV 11:32:33 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 18.355 uVs 11:32:33 Integrated Voltage Measurement... 11:32:42 Voltage Offset Correction: v_off_ini = 101.568 uV, v_off_fin = 103.802 uV 11:32:42 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -0.795 uVs 11:32:42 Integrated Voltage Measurement... 11:32:51 Voltage Offset Correction: v_off_ini = 107.397 uV, v_off_fin = 123.178 uV 11:32:51 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 25.223 uVs 11:32:53 I2X Average and RMS of Measurements VT = -21.552 +- 6.103 uVs I2X = 1.983 +- 0.562 uTm^2 11:32:58 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:33:01 Integrated Voltage Measurement... 11:33:10 Voltage Offset Correction: v_off_ini = 125.338 uV, v_off_fin = 115.013 uV 11:33:10 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = -182.642 uVs 11:33:10 Integrated Voltage Measurement... 11:33:19 Voltage Offset Correction: v_off_ini = 127.170 uV, v_off_fin = 137.838 uV 11:33:19 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 183.867 uVs 11:33:19 Integrated Voltage Measurement... 11:33:28 Voltage Offset Correction: v_off_ini = 127.134 uV, v_off_fin = 130.600 uV 11:33:28 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -163.355 uVs 11:33:28 Integrated Voltage Measurement... 11:33:37 Voltage Offset Correction: v_off_ini = 123.073 uV, v_off_fin = 124.789 uV 11:33:37 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 167.974 uVs 11:33:37 Integrated Voltage Measurement... 11:33:46 Voltage Offset Correction: v_off_ini = 115.141 uV, v_off_fin = 122.023 uV 11:33:46 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -171.049 uVs 11:33:46 Integrated Voltage Measurement... 11:33:55 Voltage Offset Correction: v_off_ini = 131.382 uV, v_off_fin = 126.742 uV 11:33:55 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 164.728 uVs 11:33:58 I2Y Average and RMS of Measurements VT = -172.269 +- 7.821 uVs I2Y = -15.849 +- 0.720 uTm^2 11:34:12 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:34:16 Integrated Voltage Measurement... 11:34:25 Voltage Offset Correction: v_off_ini = 120.507 uV, v_off_fin = 109.047 uV 11:34:25 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -70.814 uVs 11:34:25 Integrated Voltage Measurement... 11:34:34 Voltage Offset Correction: v_off_ini = 121.675 uV, v_off_fin = 116.201 uV 11:34:34 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 54.975 uVs 11:34:34 Integrated Voltage Measurement... 11:34:43 Voltage Offset Correction: v_off_ini = 109.418 uV, v_off_fin = 108.132 uV 11:34:43 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -30.526 uVs 11:34:43 Integrated Voltage Measurement... 11:34:52 Voltage Offset Correction: v_off_ini = 108.010 uV, v_off_fin = 112.671 uV 11:34:52 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 30.665 uVs 11:34:52 Integrated Voltage Measurement... 11:35:01 Voltage Offset Correction: v_off_ini = 100.123 uV, v_off_fin = 98.210 uV 11:35:01 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -40.499 uVs 11:35:01 Integrated Voltage Measurement... 11:35:10 Voltage Offset Correction: v_off_ini = 114.591 uV, v_off_fin = 123.511 uV 11:35:10 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 37.914 uVs 11:35:14 I1X Average and RMS of Measurements VT = -44.232 +- 13.656 uVs I1X = 8.847 +- 2.732 uTm 11:35:17 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:35:21 Integrated Voltage Measurement... 11:35:30 Voltage Offset Correction: v_off_ini = 122.220 uV, v_off_fin = 109.867 uV 11:35:30 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -183.282 uVs 11:35:30 Integrated Voltage Measurement... 11:35:39 Voltage Offset Correction: v_off_ini = 117.177 uV, v_off_fin = 122.723 uV 11:35:39 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 176.612 uVs 11:35:39 Integrated Voltage Measurement... 11:35:48 Voltage Offset Correction: v_off_ini = 120.341 uV, v_off_fin = 102.254 uV 11:35:48 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -180.891 uVs 11:35:48 Integrated Voltage Measurement... 11:35:57 Voltage Offset Correction: v_off_ini = 119.969 uV, v_off_fin = 132.731 uV 11:35:57 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 165.319 uVs 11:35:57 Integrated Voltage Measurement... 11:36:06 Voltage Offset Correction: v_off_ini = 114.666 uV, v_off_fin = 99.207 uV 11:36:06 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -184.490 uVs 11:36:06 Integrated Voltage Measurement... 11:36:15 Voltage Offset Correction: v_off_ini = 126.242 uV, v_off_fin = 131.759 uV 11:36:15 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 164.098 uVs 11:36:19 I1Y Average and RMS of Measurements VT = -175.782 +- 2.985 uVs I1Y = -35.156 +- 0.597 uTm 11:36:24 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:36:27 Integrated Voltage Measurement... 11:36:36 Voltage Offset Correction: v_off_ini = 99.023 uV, v_off_fin = 98.657 uV 11:36:36 Integrated Voltage Measurement dXYus = -1.0001 mm, dXYds = 0.0000 mm dXY = -1.0001 mm VT = -1.655 uVs 11:36:36 Integrated Voltage Measurement... 11:36:45 Voltage Offset Correction: v_off_ini = 111.413 uV, v_off_fin = 121.150 uV 11:36:45 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 26.948 uVs 11:36:45 Integrated Voltage Measurement... 11:36:54 Voltage Offset Correction: v_off_ini = 135.179 uV, v_off_fin = 107.215 uV 11:36:54 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -45.862 uVs 11:36:54 Integrated Voltage Measurement... 11:37:03 Voltage Offset Correction: v_off_ini = 114.979 uV, v_off_fin = 121.219 uV 11:37:03 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 24.616 uVs 11:37:03 Integrated Voltage Measurement... 11:37:12 Voltage Offset Correction: v_off_ini = 107.117 uV, v_off_fin = 114.925 uV 11:37:12 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -41.166 uVs 11:37:12 Integrated Voltage Measurement... 11:37:21 Voltage Offset Correction: v_off_ini = 117.372 uV, v_off_fin = 123.377 uV 11:37:21 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 32.416 uVs 11:37:24 I2X Average and RMS of Measurements VT = -28.777 +- 10.255 uVs I2X = 2.647 +- 0.943 uTm^2 11:37:29 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:37:31 Integrated Voltage Measurement... 11:37:40 Voltage Offset Correction: v_off_ini = 122.143 uV, v_off_fin = 121.063 uV 11:37:40 Integrated Voltage Measurement dXYus = 1.0001 mm, dXYds = 0.0000 mm dXY = 1.0001 mm VT = -149.891 uVs 11:37:40 Integrated Voltage Measurement... 11:37:49 Voltage Offset Correction: v_off_ini = 112.441 uV, v_off_fin = 114.873 uV 11:37:49 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 190.232 uVs 11:37:49 Integrated Voltage Measurement... 11:37:58 Voltage Offset Correction: v_off_ini = 110.656 uV, v_off_fin = 114.048 uV 11:37:58 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -152.591 uVs 11:37:58 Integrated Voltage Measurement... 11:38:07 Voltage Offset Correction: v_off_ini = 123.882 uV, v_off_fin = 130.013 uV 11:38:07 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 162.699 uVs 11:38:08 Integrated Voltage Measurement... 11:38:17 Voltage Offset Correction: v_off_ini = 130.142 uV, v_off_fin = 116.818 uV 11:38:17 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -173.636 uVs 11:38:17 Integrated Voltage Measurement... 11:38:26 Voltage Offset Correction: v_off_ini = 123.327 uV, v_off_fin = 140.241 uV 11:38:26 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 189.237 uVs 11:38:28 I2Y Average and RMS of Measurements VT = -169.714 +- 9.716 uVs I2Y = -15.613 +- 0.894 uTm^2 11:38:42 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:38:46 Integrated Voltage Measurement... 11:38:55 Voltage Offset Correction: v_off_ini = 113.490 uV, v_off_fin = 106.731 uV 11:38:55 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -59.157 uVs 11:38:55 Integrated Voltage Measurement... 11:39:04 Voltage Offset Correction: v_off_ini = 108.578 uV, v_off_fin = 112.789 uV 11:39:04 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 54.274 uVs 11:39:04 Integrated Voltage Measurement... 11:39:13 Voltage Offset Correction: v_off_ini = 111.157 uV, v_off_fin = 102.630 uV 11:39:13 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -49.893 uVs 11:39:13 Integrated Voltage Measurement... 11:39:22 Voltage Offset Correction: v_off_ini = 106.081 uV, v_off_fin = 119.801 uV 11:39:22 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 62.173 uVs 11:39:22 Integrated Voltage Measurement... 11:39:31 Voltage Offset Correction: v_off_ini = 117.809 uV, v_off_fin = 106.377 uV 11:39:31 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -48.348 uVs 11:39:31 Integrated Voltage Measurement... 11:39:40 Voltage Offset Correction: v_off_ini = 112.470 uV, v_off_fin = 118.057 uV 11:39:40 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 76.115 uVs 11:39:45 I1X Average and RMS of Measurements VT = -58.327 +- 2.775 uVs I1X = 11.666 +- 0.555 uTm 11:39:47 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:39:51 Integrated Voltage Measurement... 11:40:00 Voltage Offset Correction: v_off_ini = 123.297 uV, v_off_fin = 111.994 uV 11:40:00 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -186.892 uVs 11:40:00 Integrated Voltage Measurement... 11:40:09 Voltage Offset Correction: v_off_ini = 114.403 uV, v_off_fin = 124.590 uV 11:40:09 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 192.993 uVs 11:40:09 Integrated Voltage Measurement... 11:40:18 Voltage Offset Correction: v_off_ini = 107.371 uV, v_off_fin = 105.874 uV 11:40:18 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -176.296 uVs 11:40:18 Integrated Voltage Measurement... 11:40:27 Voltage Offset Correction: v_off_ini = 117.608 uV, v_off_fin = 126.592 uV 11:40:27 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 198.770 uVs 11:40:27 Integrated Voltage Measurement... 11:40:36 Voltage Offset Correction: v_off_ini = 117.669 uV, v_off_fin = 114.003 uV 11:40:36 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -148.911 uVs 11:40:36 Integrated Voltage Measurement... 11:40:45 Voltage Offset Correction: v_off_ini = 125.665 uV, v_off_fin = 140.427 uV 11:40:45 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 163.342 uVs 11:40:50 I1Y Average and RMS of Measurements VT = -177.867 +- 15.405 uVs I1Y = -35.571 +- 3.079 uTm 11:40:55 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:40:57 Integrated Voltage Measurement... 11:41:06 Voltage Offset Correction: v_off_ini = 120.428 uV, v_off_fin = 110.453 uV 11:41:06 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -33.364 uVs 11:41:06 Integrated Voltage Measurement... 11:41:15 Voltage Offset Correction: v_off_ini = 115.564 uV, v_off_fin = 134.751 uV 11:41:15 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 28.039 uVs 11:41:15 Integrated Voltage Measurement... 11:41:24 Voltage Offset Correction: v_off_ini = 146.636 uV, v_off_fin = 126.514 uV 11:41:24 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -60.897 uVs 11:41:24 Integrated Voltage Measurement... 11:41:33 Voltage Offset Correction: v_off_ini = 132.749 uV, v_off_fin = 145.561 uV 11:41:33 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 33.756 uVs 11:41:33 Integrated Voltage Measurement... 11:41:42 Voltage Offset Correction: v_off_ini = 144.623 uV, v_off_fin = 132.287 uV 11:41:43 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -59.106 uVs 11:41:43 Integrated Voltage Measurement... 11:41:52 Voltage Offset Correction: v_off_ini = 140.470 uV, v_off_fin = 146.271 uV 11:41:52 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 38.856 uVs 11:41:54 I2X Average and RMS of Measurements VT = -42.336 +- 8.255 uVs I2X = 3.895 +- 0.759 uTm^2 11:41:59 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:42:02 Integrated Voltage Measurement... 11:42:11 Voltage Offset Correction: v_off_ini = 124.339 uV, v_off_fin = 153.139 uV 11:42:11 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -173.359 uVs 11:42:11 Integrated Voltage Measurement... 11:42:20 Voltage Offset Correction: v_off_ini = 130.248 uV, v_off_fin = 140.701 uV 11:42:20 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 164.645 uVs 11:42:20 Integrated Voltage Measurement... 11:42:29 Voltage Offset Correction: v_off_ini = 120.494 uV, v_off_fin = 135.889 uV 11:42:29 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -153.117 uVs 11:42:29 Integrated Voltage Measurement... 11:42:38 Voltage Offset Correction: v_off_ini = 127.488 uV, v_off_fin = 141.571 uV 11:42:38 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 158.779 uVs 11:42:38 Integrated Voltage Measurement... 11:42:47 Voltage Offset Correction: v_off_ini = 134.125 uV, v_off_fin = 137.798 uV 11:42:47 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -174.526 uVs 11:42:47 Integrated Voltage Measurement... 11:42:56 Voltage Offset Correction: v_off_ini = 132.190 uV, v_off_fin = 143.018 uV 11:42:56 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 161.869 uVs 11:42:58 I2Y Average and RMS of Measurements VT = -164.382 +- 5.973 uVs I2Y = -15.123 +- 0.550 uTm^2 11:43:01 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:45:26 Wire Position: X0 = 50.0000 mm, Y0 = 0.0000 mm