SLAC Magnetic Measurements Date: 02-22-2018 Time: 08:54:31 Project: FACET_II Magnet Type: Dipole Magnet Name: 1.182D6.82T-1 Serial Number: 1 Run Number: 2 Measurement Device: 10wire Operator: sda Comment: Integrated field measurement at 300 A for coil calibration. Standardization for BEND: 3 cycles from 0.0 A to 300.0 A to 0.0 A Test Currents for BEND (A): 300.00 300.00 300.00 300.00 08:55:11 Position moved: 0.000000 m 08:55:14 Magnet Current for BEND, Imag = -0.0067 A Vtrans = -0.000067 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 08:55:15 Magnet Current for BEND, Imag = -0.0067 A Vtrans = -0.000067 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 08:55:15 Magnet Current for BEND, Imag = -0.0067 A Vtrans = -0.000067 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 08:55:16 Magnet Current for BEND, Imag = -0.0068 A Vtrans = -0.000068 V, Factor V/A = 0.010000 V/A, Zero Current = 0.000000 A 08:55:16 Magnet Avgerage Current Measurement for BEND : Imag = -0.0067 +- 0.0000 A 08:55:42 Initial Zero Current Measurement for BEND, Imag_zero = -0.00671347 A 08:55:42 Ambient Measurement Device(s) Being Used, type = None 08:55:42 Magnet Voltage Measurement Device(s) Being Used, type = HP34970 08:55:42 Temperature Measurement Device(s) Being Used, type = HPTHERM_HP34970 08:55:43 XMAG Device(s) Being Used, type = None 08:55:43 Ambient Measurement Device(s) Being Used, type = HP34970 08:55:43 Magnet system BEND current ramping to 300.0000 A... 09:57:18 Magnet Current for BEND, Imag = 299.9947 A Vtrans = 2.999880 V, Factor V/A = 0.010000 V/A, Zero Current = -0.006713 A 09:57:19 MAIN (V) Voltage Measurement, V = 7.564277 V 09:57:19 Temperature Measurement, Probe Number = 0, T = 30.555000 C 09:57:20 Temperature Measurement, Probe Number = 1, T = 31.503000 C 09:57:21 Temperature Measurement, Probe Number = 2, T = 16.254000 C 09:57:21 Sensor Voltage Measurement, Sensor Number = 0, V = 2.240357 V 09:57:21 Sensor Measurement, Sensor Number = 0, Sensor Type = WATER_HPLEIGHTON_T Raw Sensor Output = 2.240357 (V, A, ohm, ...), Measured Value = 27.227181 (C, atm, Hz, ...) 09:57:22 Sensor Voltage Measurement, Sensor Number = 1, V = 2.890801 V 09:57:22 Sensor Measurement, Sensor Number = 1, Sensor Type = WATER_HPLEIGHTON_T Raw Sensor Output = 2.890801 (V, A, ohm, ...), Measured Value = 36.261122 (C, atm, Hz, ...) 09:57:23 Sensor Voltage Measurement, Sensor Number = 2, V = 2.661532 V 09:57:23 Sensor Measurement, Sensor Number = 2, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 2.661532 (V, A, ohm, ...), Measured Value = 124.614877 (C, atm, Hz, ...) 09:57:23 Sensor Voltage Measurement, Sensor Number = 3, V = 1.085930 V 09:57:23 Sensor Measurement, Sensor Number = 3, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 1.085930 (V, A, ohm, ...), Measured Value = 6.444765 (C, atm, Hz, ...) 09:57:23 Magnet Test Parameter Measurements P0 = BEND, Value = 299.99 P1 = MAIN (V), Value = 7.56 P2 = Top Coil (C), Value = 30.55 P3 = Bottom Coil (C), Value = 31.50 P4 = Ambient Temp (C), Value = 16.25 P5 = Supply Temp(C), Value = 27.23 P6 = Return Temp(C), Value = 36.26 P7 = Supply Pres(psi), Value = 124.61 P8 = Return Pres(psi), Value = 6.44 09:57:23 Magnet system BEND current ramping to 0.0000 A... 09:58:58 Magnet system BEND current ramping to 300.0000 A... 10:00:32 Magnet system BEND current ramping to 0.0000 A... 10:02:06 Magnet system BEND current ramping to 300.0000 A... 10:03:41 Magnet system BEND current ramping to 0.0000 A... 10:05:15 Magnet system BEND current ramping to 300.0000 A... 10:06:49 Magnet system BEND current ramping to 0.0000 A... 10:08:26 Position moved: 0.000000 m 10:08:29 Magnet system BEND current ramping to 300.0000 A... 10:10:04 Magnet Current for BEND, Imag = 299.9921 A Vtrans = 2.999854 V, Factor V/A = 0.010000 V/A, Zero Current = -0.006713 A 10:10:05 MAIN (V) Voltage Measurement, V = 7.540509 V 10:10:05 Temperature Measurement, Probe Number = 0, T = 27.236000 C 10:10:06 Temperature Measurement, Probe Number = 1, T = 28.364000 C 10:10:06 Temperature Measurement, Probe Number = 2, T = 16.349000 C 10:10:07 Sensor Voltage Measurement, Sensor Number = 0, V = 2.236413 V 10:10:07 Sensor Measurement, Sensor Number = 0, Sensor Type = WATER_HPLEIGHTON_T Raw Sensor Output = 2.236413 (V, A, ohm, ...), Measured Value = 27.172401 (C, atm, Hz, ...) 10:10:08 Sensor Voltage Measurement, Sensor Number = 1, V = 2.575031 V 10:10:08 Sensor Measurement, Sensor Number = 1, Sensor Type = WATER_HPLEIGHTON_T Raw Sensor Output = 2.575031 (V, A, ohm, ...), Measured Value = 31.875435 (C, atm, Hz, ...) 10:10:08 Sensor Voltage Measurement, Sensor Number = 2, V = 2.661377 V 10:10:08 Sensor Measurement, Sensor Number = 2, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 2.661377 (V, A, ohm, ...), Measured Value = 124.603245 (C, atm, Hz, ...) 10:10:09 Sensor Voltage Measurement, Sensor Number = 3, V = 1.084985 V 10:10:09 Sensor Measurement, Sensor Number = 3, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 1.084985 (V, A, ohm, ...), Measured Value = 6.373845 (C, atm, Hz, ...) 10:10:09 Magnet Test Parameter Measurements P0 = BEND, Value = 299.99 P1 = MAIN (V), Value = 7.54 P2 = Top Coil (C), Value = 27.24 P3 = Bottom Coil (C), Value = 28.36 P4 = Ambient Temp (C), Value = 16.35 P5 = Supply Temp(C), Value = 27.17 P6 = Return Temp(C), Value = 31.88 P7 = Supply Pres(psi), Value = 124.60 P8 = Return Pres(psi), Value = 6.37 10:10:09 Magnetic Field Measurement Measuring the magnet strength... 10:10:09 Integrated Voltage Measurement... 10:10:10 Magnet Current for BEND, Imag = 299.9933 A Vtrans = 2.999866 V, Factor V/A = 0.010000 V/A, Zero Current = -0.006713 A 10:10:18 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0180000 m VT = 0.034252144 Vs 10:10:26 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0180000 m VT = -0.034253942 Vs 10:10:26 Magnet Current for BEND, Imag = 299.9942 A Vtrans = 2.999875 V, Factor V/A = 0.010000 V/A, Zero Current = -0.006713 A 10:10:34 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0180000 m VT = 0.034251912 Vs 10:10:41 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0180000 m VT = -0.034251618 Vs 10:10:42 Magnet Current for BEND, Imag = 299.9935 A Vtrans = 2.999868 V, Factor V/A = 0.010000 V/A, Zero Current = -0.006713 A 10:10:49 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0180000 m VT = 0.034252306 Vs 10:10:56 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0180000 m VT = -0.034246220 Vs 10:10:57 Magnet Current for BEND, Imag = 299.9954 A Vtrans = 2.999887 V, Factor V/A = 0.010000 V/A, Zero Current = -0.006713 A 10:11:04 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0180000 m VT = 0.034257871 Vs 10:11:12 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0180000 m VT = -0.034249643 Vs 10:11:12 Magnet Current for BEND, Imag = 299.9944 A Vtrans = 2.999877 V, Factor V/A = 0.010000 V/A, Zero Current = -0.006713 A 10:11:20 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0180000 m VT = 0.034256923 Vs 10:11:27 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0180000 m VT = -0.034247595 Vs 10:11:27 Integrated Voltage Average and RMS of Measurements X0 = 0.0000000 m, Delta X = 0.0180000 m VT = 0.034252018 +- 0.000001535 Vs 10:11:27 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0000000 m, dx = 0.0180000 m, Nturns = 10.0 VT = 0.034252018 +- 0.000001535 Vs BL = 0.19028899 +- 0.00000853 Tm 10:11:31 Position moved: 0.000000 m 10:11:31 BL vs X Fit Coefficients: Cn sigCn n (Tm/m^n) ----- --------+-------- 0 0.19029 0.00001 Compute Fit Residuals: X0 BL sigBL BLfit BL-BLfit (m) (Tm) (Tm) (Tm) (Tm) ----------- --------+-------- -------- -------- 0.0000000 0.19029 0.00001 0.19029 0.00000 10:11:31 Magnet Strength Calculation Harmonic Number = 1 Number Of X0 Positions = 1 X0 BL sigBL (m) (Tm) (Tm) -------- --------+-------- 0.0000 0.19029 0.00001 SL = 0.1902890 +- 0.0000085 Tm^2/m^n 10:11:31 Magnet Strength Measurement Harmonic Number = 1 SL = 0.1902890 +- 0.0000085 Tm^2/m^n 10:11:34 Position moved: 0.000000 m 10:11:34 Magnet system BEND current ramping to 300.0000 A... 10:11:45 Magnet Current for BEND, Imag = 299.9943 A Vtrans = 2.999876 V, Factor V/A = 0.010000 V/A, Zero Current = -0.006713 A 10:11:46 MAIN (V) Voltage Measurement, V = 7.561410 V 10:11:46 Temperature Measurement, Probe Number = 0, T = 28.565000 C 10:11:47 Temperature Measurement, Probe Number = 1, T = 28.643000 C 10:11:47 Temperature Measurement, Probe Number = 2, T = 16.385000 C 10:11:48 Sensor Voltage Measurement, Sensor Number = 0, V = 2.239288 V 10:11:48 Sensor Measurement, Sensor Number = 0, Sensor Type = WATER_HPLEIGHTON_T Raw Sensor Output = 2.239288 (V, A, ohm, ...), Measured Value = 27.212333 (C, atm, Hz, ...) 10:11:49 Sensor Voltage Measurement, Sensor Number = 1, V = 2.541349 V 10:11:49 Sensor Measurement, Sensor Number = 1, Sensor Type = WATER_HPLEIGHTON_T Raw Sensor Output = 2.541349 (V, A, ohm, ...), Measured Value = 31.407626 (C, atm, Hz, ...) 10:11:49 Sensor Voltage Measurement, Sensor Number = 2, V = 2.662673 V 10:11:49 Sensor Measurement, Sensor Number = 2, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 2.662673 (V, A, ohm, ...), Measured Value = 124.700483 (C, atm, Hz, ...) 10:11:50 Sensor Voltage Measurement, Sensor Number = 3, V = 1.084836 V 10:11:50 Sensor Measurement, Sensor Number = 3, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 1.084836 (V, A, ohm, ...), Measured Value = 6.362685 (C, atm, Hz, ...) 10:11:50 Magnet Test Parameter Measurements P0 = BEND, Value = 299.99 P1 = MAIN (V), Value = 7.56 P2 = Top Coil (C), Value = 28.57 P3 = Bottom Coil (C), Value = 28.64 P4 = Ambient Temp (C), Value = 16.39 P5 = Supply Temp(C), Value = 27.21 P6 = Return Temp(C), Value = 31.41 P7 = Supply Pres(psi), Value = 124.70 P8 = Return Pres(psi), Value = 6.36 10:11:50 Magnetic Field Measurement Measuring the magnet strength... 10:11:50 Integrated Voltage Measurement... 10:11:51 Magnet Current for BEND, Imag = 299.9960 A Vtrans = 2.999892 V, Factor V/A = 0.010000 V/A, Zero Current = -0.006713 A 10:11:59 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0180000 m VT = 0.034253327 Vs 10:12:06 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0180000 m VT = -0.034246529 Vs 10:12:07 Magnet Current for BEND, Imag = 299.9949 A Vtrans = 2.999882 V, Factor V/A = 0.010000 V/A, Zero Current = -0.006713 A 10:12:15 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0180000 m VT = 0.034259421 Vs 10:12:22 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0180000 m VT = -0.034253614 Vs 10:12:23 Magnet Current for BEND, Imag = 299.9960 A Vtrans = 2.999892 V, Factor V/A = 0.010000 V/A, Zero Current = -0.006713 A 10:12:30 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0180000 m VT = 0.034253044 Vs 10:12:37 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0180000 m VT = -0.034247322 Vs 10:12:38 Magnet Current for BEND, Imag = 299.9949 A Vtrans = 2.999882 V, Factor V/A = 0.010000 V/A, Zero Current = -0.006713 A 10:12:45 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0180000 m VT = 0.034255605 Vs 10:12:52 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0180000 m VT = -0.034250668 Vs 10:12:53 Magnet Current for BEND, Imag = 299.9947 A Vtrans = 2.999880 V, Factor V/A = 0.010000 V/A, Zero Current = -0.006713 A 10:13:01 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0180000 m VT = 0.034257001 Vs 10:13:08 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0180000 m VT = -0.034247516 Vs 10:13:08 Integrated Voltage Average and RMS of Measurements X0 = 0.0000000 m, Delta X = 0.0180000 m VT = 0.034252405 +- 0.000002389 Vs 10:13:08 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0000000 m, dx = 0.0180000 m, Nturns = 10.0 VT = 0.034252405 +- 0.000002389 Vs BL = 0.19029114 +- 0.00001327 Tm 10:13:12 Position moved: 0.000000 m 10:13:12 BL vs X Fit Coefficients: Cn sigCn n (Tm/m^n) ----- --------+-------- 0 0.19029 0.00001 Compute Fit Residuals: X0 BL sigBL BLfit BL-BLfit (m) (Tm) (Tm) (Tm) (Tm) ----------- --------+-------- -------- -------- 0.0000000 0.19029 0.00001 0.19029 0.00000 10:13:12 Magnet Strength Calculation Harmonic Number = 1 Number Of X0 Positions = 1 X0 BL sigBL (m) (Tm) (Tm) -------- --------+-------- 0.0000 0.19029 0.00001 SL = 0.1902911 +- 0.0000133 Tm^2/m^n 10:13:12 Magnet Strength Measurement Harmonic Number = 1 SL = 0.1902911 +- 0.0000133 Tm^2/m^n 10:13:15 Position moved: 0.000000 m 10:13:15 Magnet system BEND current ramping to 300.0000 A... 10:13:26 Magnet Current for BEND, Imag = 299.9952 A Vtrans = 2.999885 V, Factor V/A = 0.010000 V/A, Zero Current = -0.006713 A 10:13:27 MAIN (V) Voltage Measurement, V = 7.562889 V 10:13:27 Temperature Measurement, Probe Number = 0, T = 29.418000 C 10:13:28 Temperature Measurement, Probe Number = 1, T = 29.235000 C 10:13:29 Temperature Measurement, Probe Number = 2, T = 16.446000 C 10:13:29 Sensor Voltage Measurement, Sensor Number = 0, V = 2.240333 V 10:13:29 Sensor Measurement, Sensor Number = 0, Sensor Type = WATER_HPLEIGHTON_T Raw Sensor Output = 2.240333 (V, A, ohm, ...), Measured Value = 27.226844 (C, atm, Hz, ...) 10:13:30 Sensor Voltage Measurement, Sensor Number = 1, V = 2.580342 V 10:13:30 Sensor Measurement, Sensor Number = 1, Sensor Type = WATER_HPLEIGHTON_T Raw Sensor Output = 2.580342 (V, A, ohm, ...), Measured Value = 31.949192 (C, atm, Hz, ...) 10:13:31 Sensor Voltage Measurement, Sensor Number = 2, V = 2.662826 V 10:13:31 Sensor Measurement, Sensor Number = 2, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 2.662826 (V, A, ohm, ...), Measured Value = 124.711928 (C, atm, Hz, ...) 10:13:31 Sensor Voltage Measurement, Sensor Number = 3, V = 1.084489 V 10:13:31 Sensor Measurement, Sensor Number = 3, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 1.084489 (V, A, ohm, ...), Measured Value = 6.336660 (C, atm, Hz, ...) 10:13:31 Magnet Test Parameter Measurements P0 = BEND, Value = 300.00 P1 = MAIN (V), Value = 7.56 P2 = Top Coil (C), Value = 29.42 P3 = Bottom Coil (C), Value = 29.23 P4 = Ambient Temp (C), Value = 16.45 P5 = Supply Temp(C), Value = 27.23 P6 = Return Temp(C), Value = 31.95 P7 = Supply Pres(psi), Value = 124.71 P8 = Return Pres(psi), Value = 6.34 10:13:31 Magnetic Field Measurement Measuring the magnet strength... 10:13:31 Integrated Voltage Measurement... 10:13:32 Magnet Current for BEND, Imag = 299.9939 A Vtrans = 2.999872 V, Factor V/A = 0.010000 V/A, Zero Current = -0.006713 A 10:13:41 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0180000 m VT = 0.034254512 Vs 10:13:48 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0180000 m VT = -0.034252189 Vs 10:13:49 Magnet Current for BEND, Imag = 299.9954 A Vtrans = 2.999887 V, Factor V/A = 0.010000 V/A, Zero Current = -0.006713 A 10:13:56 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0180000 m VT = 0.034254444 Vs 10:14:03 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0180000 m VT = -0.034249414 Vs 10:14:04 Magnet Current for BEND, Imag = 299.9956 A Vtrans = 2.999889 V, Factor V/A = 0.010000 V/A, Zero Current = -0.006713 A 10:14:11 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0180000 m VT = 0.034256424 Vs 10:14:19 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0180000 m VT = -0.034246854 Vs 10:14:19 Magnet Current for BEND, Imag = 299.9960 A Vtrans = 2.999892 V, Factor V/A = 0.010000 V/A, Zero Current = -0.006713 A 10:14:27 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0180000 m VT = 0.034258878 Vs 10:14:34 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0180000 m VT = -0.034254225 Vs 10:14:35 Magnet Current for BEND, Imag = 299.9940 A Vtrans = 2.999873 V, Factor V/A = 0.010000 V/A, Zero Current = -0.006713 A 10:14:42 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0180000 m VT = 0.034254428 Vs 10:14:50 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0180000 m VT = -0.034246275 Vs 10:14:50 Integrated Voltage Average and RMS of Measurements X0 = 0.0000000 m, Delta X = 0.0180000 m VT = 0.034252764 +- 0.000002120 Vs 10:14:50 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0000000 m, dx = 0.0180000 m, Nturns = 10.0 VT = 0.034252764 +- 0.000002120 Vs BL = 0.19029314 +- 0.00001178 Tm 10:14:54 Position moved: 0.000000 m 10:14:54 BL vs X Fit Coefficients: Cn sigCn n (Tm/m^n) ----- --------+-------- 0 0.19029 0.00001 Compute Fit Residuals: X0 BL sigBL BLfit BL-BLfit (m) (Tm) (Tm) (Tm) (Tm) ----------- --------+-------- -------- -------- 0.0000000 0.19029 0.00001 0.19029 0.00000 10:14:54 Magnet Strength Calculation Harmonic Number = 1 Number Of X0 Positions = 1 X0 BL sigBL (m) (Tm) (Tm) -------- --------+-------- 0.0000 0.19029 0.00001 SL = 0.1902931 +- 0.0000118 Tm^2/m^n 10:14:54 Magnet Strength Measurement Harmonic Number = 1 SL = 0.1902931 +- 0.0000118 Tm^2/m^n 10:14:57 Position moved: 0.000000 m 10:14:57 Magnet system BEND current ramping to 300.0000 A... 10:15:08 Magnet Current for BEND, Imag = 299.9947 A Vtrans = 2.999880 V, Factor V/A = 0.010000 V/A, Zero Current = -0.006713 A 10:15:08 MAIN (V) Voltage Measurement, V = 7.563549 V 10:15:09 Temperature Measurement, Probe Number = 0, T = 29.989000 C 10:15:10 Temperature Measurement, Probe Number = 1, T = 29.912000 C 10:15:10 Temperature Measurement, Probe Number = 2, T = 16.441000 C 10:15:11 Sensor Voltage Measurement, Sensor Number = 0, V = 2.241069 V 10:15:11 Sensor Measurement, Sensor Number = 0, Sensor Type = WATER_HPLEIGHTON_T Raw Sensor Output = 2.241069 (V, A, ohm, ...), Measured Value = 27.237067 (C, atm, Hz, ...) 10:15:11 Sensor Voltage Measurement, Sensor Number = 1, V = 2.654635 V 10:15:12 Sensor Measurement, Sensor Number = 1, Sensor Type = WATER_HPLEIGHTON_T Raw Sensor Output = 2.654635 (V, A, ohm, ...), Measured Value = 32.981040 (C, atm, Hz, ...) 10:15:12 Sensor Voltage Measurement, Sensor Number = 2, V = 2.666306 V 10:15:12 Sensor Measurement, Sensor Number = 2, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 2.666306 (V, A, ohm, ...), Measured Value = 124.972942 (C, atm, Hz, ...) 10:15:13 Sensor Voltage Measurement, Sensor Number = 3, V = 1.084996 V 10:15:13 Sensor Measurement, Sensor Number = 3, Sensor Type = WATER_HPASHCROFT_P Raw Sensor Output = 1.084996 (V, A, ohm, ...), Measured Value = 6.374700 (C, atm, Hz, ...) 10:15:13 Magnet Test Parameter Measurements P0 = BEND, Value = 299.99 P1 = MAIN (V), Value = 7.56 P2 = Top Coil (C), Value = 29.99 P3 = Bottom Coil (C), Value = 29.91 P4 = Ambient Temp (C), Value = 16.44 P5 = Supply Temp(C), Value = 27.24 P6 = Return Temp(C), Value = 32.98 P7 = Supply Pres(psi), Value = 124.97 P8 = Return Pres(psi), Value = 6.37 10:15:13 Magnetic Field Measurement Measuring the magnet strength... 10:15:13 Integrated Voltage Measurement... 10:15:14 Magnet Current for BEND, Imag = 299.9960 A Vtrans = 2.999892 V, Factor V/A = 0.010000 V/A, Zero Current = -0.006713 A 10:15:22 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0180000 m VT = 0.034252376 Vs 10:15:29 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0180000 m VT = -0.034250667 Vs 10:15:30 Magnet Current for BEND, Imag = 299.9963 A Vtrans = 2.999896 V, Factor V/A = 0.010000 V/A, Zero Current = -0.006713 A 10:15:37 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0180000 m VT = 0.034253267 Vs 10:15:45 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0180000 m VT = -0.034253626 Vs 10:15:45 Magnet Current for BEND, Imag = 299.9954 A Vtrans = 2.999887 V, Factor V/A = 0.010000 V/A, Zero Current = -0.006713 A 10:15:53 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0180000 m VT = 0.034256950 Vs 10:16:00 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0180000 m VT = -0.034238733 Vs 10:16:01 Magnet Current for BEND, Imag = 299.9963 A Vtrans = 2.999896 V, Factor V/A = 0.010000 V/A, Zero Current = -0.006713 A 10:16:08 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0180000 m VT = 0.034259098 Vs 10:16:15 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0180000 m VT = -0.034249744 Vs 10:16:16 Magnet Current for BEND, Imag = 299.9960 A Vtrans = 2.999892 V, Factor V/A = 0.010000 V/A, Zero Current = -0.006713 A 10:16:24 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = 0.0180000 m VT = 0.034257980 Vs 10:16:31 Integrated Voltage Measurement X0 = 0.0000000 m, Delta X = -0.0180000 m VT = -0.034255595 Vs 10:16:31 Integrated Voltage Average and RMS of Measurements X0 = 0.0000000 m, Delta X = 0.0180000 m VT = 0.034252804 +- 0.000003006 Vs 10:16:31 Integrated Field Strength, Average and RMS of Measurements X0 = 0.0000000 m, dx = 0.0180000 m, Nturns = 10.0 VT = 0.034252804 +- 0.000003006 Vs BL = 0.19029335 +- 0.00001670 Tm 10:16:35 Position moved: 0.000000 m 10:16:35 BL vs X Fit Coefficients: Cn sigCn n (Tm/m^n) ----- --------+-------- 0 0.19029 0.00002 Compute Fit Residuals: X0 BL sigBL BLfit BL-BLfit (m) (Tm) (Tm) (Tm) (Tm) ----------- --------+-------- -------- -------- 0.0000000 0.19029 0.00002 0.19029 0.00000 10:16:35 Magnet Strength Calculation Harmonic Number = 1 Number Of X0 Positions = 1 X0 BL sigBL (m) (Tm) (Tm) -------- --------+-------- 0.0000 0.19029 0.00002 SL = 0.1902934 +- 0.0000167 Tm^2/m^n 10:16:35 Magnet Strength Measurement Harmonic Number = 1 SL = 0.1902934 +- 0.0000167 Tm^2/m^n 10:16:38 Position moved: 0.000000 m 10:16:41 Position moved: 0.000000 m 10:16:43 Magnet system BEND current ramping to 0.0000 A...