SLAC Magnetic Measurements Date: 03-26-2019 Time: 10:39:24 Project: LCLS-II Device Type: Undulator Data Type: Serial Number: HXU_002 DATASET Number: 1 Tuning Stage: Tuning Run Number: 5 Measurement Type: Long Coil Scans Operator: VV Comment: pole #116 stronger with 3 small slags Type Position Raw value Integral Nominal Actual Pre-measurement Gap Actual Post-measurement Gap m Volt*sec T*m | T*m^2 Gap US - mm US Raw DS - mm DS Raw US - mm US Raw DS - mm DS Raw --------------------------------------------------------------------------------------------------------------------------------------- I1X_VS_X 0.000E-03 2.708E-06 60.163E-06 9.00000 8.99965 4.49962 8.99980 4.49940 8.99960 4.49972 8.99980 4.49940 I1X_VS_X 0.000E-03 2.770E-06 61.543E-06 9.00000 8.99960 4.49972 8.99975 4.49940 8.99960 4.49962 8.99980 4.49930 I1X_VS_X 0.000E-03 2.595E-06 57.653E-06 9.00000 8.99960 4.49972 8.99975 4.49940 8.99955 4.49962 8.99975 4.49940