Values of wire location in magnet and Roll, Pitch and Yaw of magnet Run X Y Roll Pitch Yaw (mm) (mm) (mrad) (mrad) (mrad) 8 333.0528 262.7614 0.5879 -0.3465 -6.3851 7 333.0559 262.7623 0.5739 -0.2475 -6.8432 Mean 333.0544 262.7618 0.5809 -0.2970 -6.6141 Std 0.0022 0.0006 0.0099 0.0700 0.3240 Error 0.00371 0.00164 Difference of wire location from Wire Detector and End Pins Run PD1X Diff PD2Y Diff PD3Y Diff PD4X Diff MagnetX Diff MagnetY Diff (microns) (microns) (microns) (microns) (microns) (microns) 8 -51.1469 6.7973 5.1577 -9.2732 -30.5693 7.5063 7 -46.8570 6.9574 6.7677 -7.3731 -27.4547 8.3767 Mean -49.0020 6.8773 5.9627 -8.3231 -29.0120 7.9415 Std 3.0334 0.1132 1.1384 1.3435 2.2024 0.6155