SLAC Magnetic Measurements Date: 10-04-2022 Time: 13:22:40 Project: LCLS-II-HE Device Type: Undulator Data Type: Serial Number: HE_SXU_000 DATASET Number: 1 Tuning Stage: Tuning Run Number: 76 Measurement Type: Long Coil Scans Operator: YL Comment: Skew Quad, 4V Center position: X0=0.002000, Y0=0.000000 Nominal Gap=12.00 Scan Raw value Integral Nominal Actual Pre-measurement Gap Actual Post-measurement Gap Number Volt*sec T*m | T*m^2 Gap US - mm US Raw DS - mm DS Raw US - mm US Raw DS - mm DS Raw --------------------------------------------------------------------------------------------------------------------------------------- 0 0.831E-06 11.863E-06 12.00000 12.00045 0 12.00005 0 12.00045 0 12.00005 0 1 0.619E-06 8.841E-06 12.00000 12.00045 0 12.00005 0 12.00045 0 12.00005 0 2 0.718E-06 10.253E-06 12.00000 12.00040 0 12.00005 0 12.00045 0 12.00005 0 3 0.595E-06 8.496E-06 12.00000 12.00045 0 12.00005 0 12.00040 0 12.00005 0 4 0.669E-06 9.551E-06 12.00000 12.00045 0 12.00005 0 12.00045 0 12.00005 0