SLAC Magnetic Measurements Date: 08-13-2026 Time: 08:25:01 Project: LCLS-II-HE Magnet Type: Phase Shifter Magnet Name: PS_DELTA-II Dataset Number: 1 Run Purpose: Tuning Run Type: Run Number: 13 Measurement Device: Stretched Wire Operator: YL Comment: 08:25:10 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 08:25:12 Motion Parameters, X Controller, Axis 1 acc = 0.001000 m/s/s dcl = 0.001000 m/s/s vel = 0.001000 m/s pos = 0.000000 m kp = 188.600000 ki = 820.000000 kd = 491.400000 db = 0.000000 08:25:13 Motion Parameters, X Controller, Axis 2 acc = 0.001000 m/s/s dcl = 0.001000 m/s/s vel = 0.001000 m/s pos = 0.000000 m kp = 188.600000 ki = 820.000000 kd = 491.400000 db = 0.000000 08:25:15 Motion Parameters, Y Controller, Axis 1 acc = 0.001000 m/s/s dcl = 0.001000 m/s/s vel = 0.001000 m/s pos = 0.000000 m kp = 188.600000 ki = 820.000000 kd = 491.400000 db = 0.000000 08:25:17 Motion Parameters, Y Controller, Axis 2 acc = 0.001000 m/s/s dcl = 0.001000 m/s/s vel = 0.001000 m/s pos = 0.000000 m kp = 188.600000 ki = 820.000000 kd = 491.400000 db = 0.000000 08:25:41 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 08:25:46 Integrated Voltage Measurement... 08:25:55 Voltage Offset Correction: v_off_ini = 2866.794 uV, v_off_fin = 2856.997 uV 08:25:55 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -56.682 uVs 08:25:55 Integrated Voltage Measurement... 08:26:04 Voltage Offset Correction: v_off_ini = 2852.951 uV, v_off_fin = 2845.750 uV 08:26:04 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 40.895 uVs 08:26:04 Integrated Voltage Measurement... 08:26:13 Voltage Offset Correction: v_off_ini = 2853.435 uV, v_off_fin = 2847.175 uV 08:26:13 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -88.719 uVs 08:26:13 Integrated Voltage Measurement... 08:26:22 Voltage Offset Correction: v_off_ini = 2855.419 uV, v_off_fin = 2857.721 uV 08:26:22 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 76.546 uVs 08:26:22 Integrated Voltage Measurement... 08:26:31 Voltage Offset Correction: v_off_ini = 2860.282 uV, v_off_fin = 2839.741 uV 08:26:31 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -58.510 uVs 08:26:31 Integrated Voltage Measurement... 08:26:40 Voltage Offset Correction: v_off_ini = 2857.131 uV, v_off_fin = 2857.204 uV 08:26:40 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 75.454 uVs 08:26:44 I1X Average and RMS of Measurements VT = -66.134 +- 13.830 uVs I1X = 13.227 +- 2.766 uTm 08:26:46 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 08:26:51 Integrated Voltage Measurement... 08:27:00 Voltage Offset Correction: v_off_ini = 2855.947 uV, v_off_fin = 2818.508 uV 08:27:00 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -205.719 uVs 08:27:00 Integrated Voltage Measurement... 08:27:09 Voltage Offset Correction: v_off_ini = 2848.832 uV, v_off_fin = 2852.081 uV 08:27:09 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 178.188 uVs 08:27:09 Integrated Voltage Measurement... 08:27:18 Voltage Offset Correction: v_off_ini = 2839.437 uV, v_off_fin = 2834.223 uV 08:27:18 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -201.401 uVs 08:27:18 Integrated Voltage Measurement... 08:27:27 Voltage Offset Correction: v_off_ini = 2839.809 uV, v_off_fin = 2863.800 uV 08:27:27 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 176.282 uVs 08:27:27 Integrated Voltage Measurement... 08:27:36 Voltage Offset Correction: v_off_ini = 2853.812 uV, v_off_fin = 2836.563 uV 08:27:36 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -189.385 uVs 08:27:36 Integrated Voltage Measurement... 08:27:45 Voltage Offset Correction: v_off_ini = 2848.618 uV, v_off_fin = 2863.908 uV 08:27:45 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 175.978 uVs 08:27:49 I1Y Average and RMS of Measurements VT = -187.826 +- 3.853 uVs I1Y = -37.565 +- 0.771 uTm 08:27:54 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 08:27:56 Integrated Voltage Measurement... 08:28:05 Voltage Offset Correction: v_off_ini = 2819.528 uV, v_off_fin = 2813.297 uV 08:28:05 Integrated Voltage Measurement dXYus = -1.0001 mm, dXYds = 0.0000 mm dXY = -1.0001 mm VT = -53.298 uVs 08:28:05 Integrated Voltage Measurement... 08:28:14 Voltage Offset Correction: v_off_ini = 2818.918 uV, v_off_fin = 2835.105 uV 08:28:14 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 66.270 uVs 08:28:15 Integrated Voltage Measurement... 08:28:23 Voltage Offset Correction: v_off_ini = 2814.882 uV, v_off_fin = 2827.245 uV 08:28:24 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -50.153 uVs 08:28:24 Integrated Voltage Measurement... 08:28:33 Voltage Offset Correction: v_off_ini = 2828.878 uV, v_off_fin = 2842.918 uV 08:28:33 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 46.549 uVs 08:28:33 Integrated Voltage Measurement... 08:28:42 Voltage Offset Correction: v_off_ini = 2818.386 uV, v_off_fin = 2805.891 uV 08:28:42 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -68.822 uVs 08:28:42 Integrated Voltage Measurement... 08:28:51 Voltage Offset Correction: v_off_ini = 2814.372 uV, v_off_fin = 2833.383 uV 08:28:51 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 66.348 uVs 08:28:53 I2X Average and RMS of Measurements VT = -58.573 +- 7.899 uVs I2X = 5.389 +- 0.727 uTm^2 08:28:58 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 08:29:01 Integrated Voltage Measurement... 08:29:10 Voltage Offset Correction: v_off_ini = 2835.924 uV, v_off_fin = 2837.711 uV 08:29:10 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -226.865 uVs 08:29:10 Integrated Voltage Measurement... 08:29:19 Voltage Offset Correction: v_off_ini = 2842.194 uV, v_off_fin = 2838.879 uV 08:29:19 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 245.466 uVs 08:29:19 Integrated Voltage Measurement... 08:29:28 Voltage Offset Correction: v_off_ini = 2835.801 uV, v_off_fin = 2826.910 uV 08:29:28 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -240.954 uVs 08:29:28 Integrated Voltage Measurement... 08:29:37 Voltage Offset Correction: v_off_ini = 2844.202 uV, v_off_fin = 2847.068 uV 08:29:37 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 245.685 uVs 08:29:37 Integrated Voltage Measurement... 08:29:46 Voltage Offset Correction: v_off_ini = 2838.713 uV, v_off_fin = 2845.515 uV 08:29:46 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -243.844 uVs 08:29:46 Integrated Voltage Measurement... 08:29:55 Voltage Offset Correction: v_off_ini = 2834.736 uV, v_off_fin = 2836.599 uV 08:29:55 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 246.284 uVs 08:29:57 I2Y Average and RMS of Measurements VT = -241.516 +- 3.850 uVs I2Y = -22.220 +- 0.354 uTm^2 08:30:08 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 08:30:12 Integrated Voltage Measurement... 08:30:21 Voltage Offset Correction: v_off_ini = 2842.288 uV, v_off_fin = 2829.757 uV 08:30:21 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -74.354 uVs 08:30:21 Integrated Voltage Measurement... 08:30:30 Voltage Offset Correction: v_off_ini = 2839.178 uV, v_off_fin = 2865.496 uV 08:30:30 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 63.439 uVs 08:30:30 Integrated Voltage Measurement... 08:30:39 Voltage Offset Correction: v_off_ini = 2851.392 uV, v_off_fin = 2842.779 uV 08:30:39 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -78.981 uVs 08:30:39 Integrated Voltage Measurement... 08:30:48 Voltage Offset Correction: v_off_ini = 2859.439 uV, v_off_fin = 2852.585 uV 08:30:48 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 86.313 uVs 08:30:48 Integrated Voltage Measurement... 08:30:57 Voltage Offset Correction: v_off_ini = 2856.348 uV, v_off_fin = 2861.396 uV 08:30:57 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -67.489 uVs 08:30:57 Integrated Voltage Measurement... 08:31:06 Voltage Offset Correction: v_off_ini = 2862.548 uV, v_off_fin = 2874.465 uV 08:31:06 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 42.173 uVs 08:31:10 I1X Average and RMS of Measurements VT = -68.792 +- 11.356 uVs I1X = 13.759 +- 2.271 uTm 08:31:13 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 08:31:17 Integrated Voltage Measurement... 08:31:26 Voltage Offset Correction: v_off_ini = 2868.321 uV, v_off_fin = 2860.682 uV 08:31:26 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -201.085 uVs 08:31:26 Integrated Voltage Measurement... 08:31:35 Voltage Offset Correction: v_off_ini = 2876.383 uV, v_off_fin = 2876.834 uV 08:31:35 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 183.077 uVs 08:31:35 Integrated Voltage Measurement... 08:31:44 Voltage Offset Correction: v_off_ini = 2876.809 uV, v_off_fin = 2867.890 uV 08:31:44 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -191.787 uVs 08:31:44 Integrated Voltage Measurement... 08:31:53 Voltage Offset Correction: v_off_ini = 2874.216 uV, v_off_fin = 2889.394 uV 08:31:53 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 199.183 uVs 08:31:53 Integrated Voltage Measurement... 08:32:02 Voltage Offset Correction: v_off_ini = 2868.109 uV, v_off_fin = 2856.858 uV 08:32:02 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -209.112 uVs 08:32:02 Integrated Voltage Measurement... 08:32:11 Voltage Offset Correction: v_off_ini = 2860.903 uV, v_off_fin = 2876.959 uV 08:32:11 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 203.134 uVs 08:32:15 I1Y Average and RMS of Measurements VT = -197.896 +- 5.981 uVs I1Y = -39.578 +- 1.194 uTm 08:32:20 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 08:32:23 Integrated Voltage Measurement... 08:32:32 Voltage Offset Correction: v_off_ini = 2870.085 uV, v_off_fin = 2872.809 uV 08:32:32 Integrated Voltage Measurement dXYus = -1.0001 mm, dXYds = 0.0000 mm dXY = -1.0001 mm VT = -79.391 uVs 08:32:32 Integrated Voltage Measurement... 08:32:41 Voltage Offset Correction: v_off_ini = 2866.542 uV, v_off_fin = 2873.369 uV 08:32:41 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 79.619 uVs 08:32:41 Integrated Voltage Measurement... 08:32:50 Voltage Offset Correction: v_off_ini = 2856.022 uV, v_off_fin = 2866.617 uV 08:32:50 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -47.966 uVs 08:32:50 Integrated Voltage Measurement... 08:32:59 Voltage Offset Correction: v_off_ini = 2895.310 uV, v_off_fin = 2889.730 uV 08:32:59 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 59.844 uVs 08:32:59 Integrated Voltage Measurement... 08:33:08 Voltage Offset Correction: v_off_ini = 2870.432 uV, v_off_fin = 2872.241 uV 08:33:08 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -84.962 uVs 08:33:08 Integrated Voltage Measurement... 08:33:17 Voltage Offset Correction: v_off_ini = 2876.975 uV, v_off_fin = 2874.481 uV 08:33:17 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 59.264 uVs 08:33:19 I2X Average and RMS of Measurements VT = -68.508 +- 10.758 uVs I2X = 6.303 +- 0.990 uTm^2 08:33:24 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 08:33:27 Integrated Voltage Measurement... 08:33:36 Voltage Offset Correction: v_off_ini = 2895.868 uV, v_off_fin = 2893.282 uV 08:33:36 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -263.683 uVs 08:33:36 Integrated Voltage Measurement... 08:33:45 Voltage Offset Correction: v_off_ini = 2875.481 uV, v_off_fin = 2881.662 uV 08:33:45 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 245.024 uVs 08:33:45 Integrated Voltage Measurement... 08:33:54 Voltage Offset Correction: v_off_ini = 2852.476 uV, v_off_fin = 2856.679 uV 08:33:54 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -235.770 uVs 08:33:54 Integrated Voltage Measurement... 08:34:03 Voltage Offset Correction: v_off_ini = 2875.276 uV, v_off_fin = 2879.330 uV 08:34:03 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 250.272 uVs 08:34:03 Integrated Voltage Measurement... 08:34:12 Voltage Offset Correction: v_off_ini = 2872.729 uV, v_off_fin = 2863.556 uV 08:34:12 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -269.867 uVs 08:34:12 Integrated Voltage Measurement... 08:34:21 Voltage Offset Correction: v_off_ini = 2862.503 uV, v_off_fin = 2868.765 uV 08:34:21 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 241.848 uVs 08:34:24 I2Y Average and RMS of Measurements VT = -251.077 +- 5.729 uVs I2Y = -23.099 +- 0.527 uTm^2 08:34:34 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 08:34:39 Integrated Voltage Measurement... 08:34:48 Voltage Offset Correction: v_off_ini = 2871.873 uV, v_off_fin = 2868.525 uV 08:34:48 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -86.398 uVs 08:34:48 Integrated Voltage Measurement... 08:34:57 Voltage Offset Correction: v_off_ini = 2865.192 uV, v_off_fin = 2868.452 uV 08:34:57 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 74.295 uVs 08:34:57 Integrated Voltage Measurement... 08:35:06 Voltage Offset Correction: v_off_ini = 2859.451 uV, v_off_fin = 2854.648 uV 08:35:06 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -68.825 uVs 08:35:06 Integrated Voltage Measurement... 08:35:15 Voltage Offset Correction: v_off_ini = 2857.722 uV, v_off_fin = 2865.044 uV 08:35:15 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 120.111 uVs 08:35:15 Integrated Voltage Measurement... 08:35:24 Voltage Offset Correction: v_off_ini = 2883.763 uV, v_off_fin = 2885.158 uV 08:35:24 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -88.911 uVs 08:35:24 Integrated Voltage Measurement... 08:35:33 Voltage Offset Correction: v_off_ini = 2854.638 uV, v_off_fin = 2881.279 uV 08:35:33 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 87.208 uVs 08:35:37 I1X Average and RMS of Measurements VT = -87.624 +- 5.773 uVs I1X = 17.526 +- 1.154 uTm 08:35:39 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 08:35:44 Integrated Voltage Measurement... 08:35:53 Voltage Offset Correction: v_off_ini = 2883.851 uV, v_off_fin = 2882.308 uV 08:35:53 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -209.484 uVs 08:35:53 Integrated Voltage Measurement... 08:36:02 Voltage Offset Correction: v_off_ini = 2874.065 uV, v_off_fin = 2889.389 uV 08:36:02 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 208.146 uVs 08:36:02 Integrated Voltage Measurement... 08:36:10 Voltage Offset Correction: v_off_ini = 2886.398 uV, v_off_fin = 2882.748 uV 08:36:11 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -189.180 uVs 08:36:11 Integrated Voltage Measurement... 08:36:19 Voltage Offset Correction: v_off_ini = 2870.201 uV, v_off_fin = 2873.228 uV 08:36:19 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 208.702 uVs 08:36:20 Integrated Voltage Measurement... 08:36:28 Voltage Offset Correction: v_off_ini = 2879.430 uV, v_off_fin = 2852.559 uV 08:36:28 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -206.764 uVs 08:36:29 Integrated Voltage Measurement... 08:36:37 Voltage Offset Correction: v_off_ini = 2872.225 uV, v_off_fin = 2872.155 uV 08:36:37 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 207.480 uVs 08:36:42 I1Y Average and RMS of Measurements VT = -204.959 +- 4.311 uVs I1Y = -40.990 +- 0.862 uTm 08:36:47 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 08:36:49 Integrated Voltage Measurement... 08:36:58 Voltage Offset Correction: v_off_ini = 2852.883 uV, v_off_fin = 2849.991 uV 08:36:58 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -64.776 uVs 08:36:58 Integrated Voltage Measurement... 08:37:07 Voltage Offset Correction: v_off_ini = 2856.430 uV, v_off_fin = 2874.268 uV 08:37:07 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 74.223 uVs 08:37:07 Integrated Voltage Measurement... 08:37:16 Voltage Offset Correction: v_off_ini = 2872.653 uV, v_off_fin = 2881.068 uV 08:37:16 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -77.986 uVs 08:37:16 Integrated Voltage Measurement... 08:37:25 Voltage Offset Correction: v_off_ini = 2898.899 uV, v_off_fin = 2907.355 uV 08:37:25 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 95.496 uVs 08:37:25 Integrated Voltage Measurement... 08:37:34 Voltage Offset Correction: v_off_ini = 2895.341 uV, v_off_fin = 2889.557 uV 08:37:34 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -81.911 uVs 08:37:34 Integrated Voltage Measurement... 08:37:43 Voltage Offset Correction: v_off_ini = 2885.598 uV, v_off_fin = 2875.960 uV 08:37:43 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 90.280 uVs 08:37:46 I2X Average and RMS of Measurements VT = -80.779 +- 7.980 uVs I2X = 7.432 +- 0.734 uTm^2 08:37:51 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 08:37:54 Integrated Voltage Measurement... 08:38:03 Voltage Offset Correction: v_off_ini = 2880.237 uV, v_off_fin = 2870.860 uV 08:38:03 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0001 mm dXY = 1.0000 mm VT = -247.993 uVs 08:38:03 Integrated Voltage Measurement... 08:38:12 Voltage Offset Correction: v_off_ini = 2870.064 uV, v_off_fin = 2880.531 uV 08:38:12 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = -0.0001 mm dXY = -1.0000 mm VT = 267.221 uVs 08:38:12 Integrated Voltage Measurement... 08:38:21 Voltage Offset Correction: v_off_ini = 2873.926 uV, v_off_fin = 2876.201 uV 08:38:21 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -274.383 uVs 08:38:21 Integrated Voltage Measurement... 08:38:30 Voltage Offset Correction: v_off_ini = 2882.864 uV, v_off_fin = 2879.985 uV 08:38:30 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 253.864 uVs 08:38:30 Integrated Voltage Measurement... 08:38:39 Voltage Offset Correction: v_off_ini = 2848.553 uV, v_off_fin = 2841.680 uV 08:38:39 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -268.971 uVs 08:38:39 Integrated Voltage Measurement... 08:38:48 Voltage Offset Correction: v_off_ini = 2843.461 uV, v_off_fin = 2859.917 uV 08:38:48 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 266.601 uVs 08:38:50 I2Y Average and RMS of Measurements VT = -263.172 +- 4.209 uVs I2Y = -24.212 +- 0.387 uTm^2 08:39:01 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 08:39:05 Integrated Voltage Measurement... 08:39:14 Voltage Offset Correction: v_off_ini = 2866.295 uV, v_off_fin = 2874.439 uV 08:39:14 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -101.772 uVs 08:39:14 Integrated Voltage Measurement... 08:39:23 Voltage Offset Correction: v_off_ini = 2882.743 uV, v_off_fin = 2886.734 uV 08:39:23 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 82.316 uVs 08:39:23 Integrated Voltage Measurement... 08:39:32 Voltage Offset Correction: v_off_ini = 2868.084 uV, v_off_fin = 2891.302 uV 08:39:32 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -104.129 uVs 08:39:32 Integrated Voltage Measurement... 08:39:41 Voltage Offset Correction: v_off_ini = 2889.666 uV, v_off_fin = 2887.405 uV 08:39:41 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 98.031 uVs 08:39:41 Integrated Voltage Measurement... 08:39:50 Voltage Offset Correction: v_off_ini = 2881.929 uV, v_off_fin = 2874.823 uV 08:39:50 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -87.476 uVs 08:39:50 Integrated Voltage Measurement... 08:39:59 Voltage Offset Correction: v_off_ini = 2870.664 uV, v_off_fin = 2878.859 uV 08:39:59 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 84.926 uVs 08:40:03 I1X Average and RMS of Measurements VT = -93.108 +- 6.121 uVs I1X = 18.622 +- 1.224 uTm 08:40:06 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 08:40:10 Integrated Voltage Measurement... 08:40:19 Voltage Offset Correction: v_off_ini = 2865.095 uV, v_off_fin = 2862.343 uV 08:40:19 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -201.639 uVs 08:40:19 Integrated Voltage Measurement... 08:40:28 Voltage Offset Correction: v_off_ini = 2883.112 uV, v_off_fin = 2892.139 uV 08:40:28 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 235.873 uVs 08:40:28 Integrated Voltage Measurement... 08:40:37 Voltage Offset Correction: v_off_ini = 2867.137 uV, v_off_fin = 2885.870 uV 08:40:37 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -232.539 uVs 08:40:37 Integrated Voltage Measurement... 08:40:46 Voltage Offset Correction: v_off_ini = 2881.231 uV, v_off_fin = 2915.611 uV 08:40:46 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 226.603 uVs 08:40:46 Integrated Voltage Measurement... 08:40:55 Voltage Offset Correction: v_off_ini = 2897.994 uV, v_off_fin = 2891.839 uV 08:40:55 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -199.671 uVs 08:40:55 Integrated Voltage Measurement... 08:41:04 Voltage Offset Correction: v_off_ini = 2880.281 uV, v_off_fin = 2904.019 uV 08:41:04 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 206.316 uVs 08:41:09 I1Y Average and RMS of Measurements VT = -217.107 +- 10.913 uVs I1Y = -43.420 +- 2.182 uTm 08:41:14 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 08:41:17 Integrated Voltage Measurement... 08:41:26 Voltage Offset Correction: v_off_ini = 2894.359 uV, v_off_fin = 2901.716 uV 08:41:26 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -110.086 uVs 08:41:26 Integrated Voltage Measurement... 08:41:35 Voltage Offset Correction: v_off_ini = 2920.427 uV, v_off_fin = 2931.561 uV 08:41:35 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 65.966 uVs 08:41:35 Integrated Voltage Measurement... 08:41:44 Voltage Offset Correction: v_off_ini = 2904.663 uV, v_off_fin = 2894.769 uV 08:41:44 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -67.710 uVs 08:41:44 Integrated Voltage Measurement... 08:41:53 Voltage Offset Correction: v_off_ini = 2896.535 uV, v_off_fin = 2911.165 uV 08:41:53 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 97.711 uVs 08:41:53 Integrated Voltage Measurement... 08:42:03 Voltage Offset Correction: v_off_ini = 2887.590 uV, v_off_fin = 2881.404 uV 08:42:03 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -76.472 uVs 08:42:03 Integrated Voltage Measurement... 08:42:12 Voltage Offset Correction: v_off_ini = 2902.558 uV, v_off_fin = 2919.859 uV 08:42:12 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 84.277 uVs 08:42:14 I2X Average and RMS of Measurements VT = -83.704 +- 3.202 uVs I2X = 7.701 +- 0.295 uTm^2 08:42:19 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 08:42:22 Integrated Voltage Measurement... 08:42:31 Voltage Offset Correction: v_off_ini = 2929.694 uV, v_off_fin = 2927.003 uV 08:42:31 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -273.674 uVs 08:42:31 Integrated Voltage Measurement... 08:42:40 Voltage Offset Correction: v_off_ini = 2924.138 uV, v_off_fin = 2932.956 uV 08:42:40 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 268.014 uVs 08:42:40 Integrated Voltage Measurement... 08:42:49 Voltage Offset Correction: v_off_ini = 2929.353 uV, v_off_fin = 2919.488 uV 08:42:49 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -250.383 uVs 08:42:49 Integrated Voltage Measurement... 08:42:58 Voltage Offset Correction: v_off_ini = 2925.772 uV, v_off_fin = 2933.838 uV 08:42:58 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 260.087 uVs 08:42:59 Integrated Voltage Measurement... 08:43:08 Voltage Offset Correction: v_off_ini = 2952.739 uV, v_off_fin = 2937.005 uV 08:43:08 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -269.039 uVs 08:43:08 Integrated Voltage Measurement... 08:43:17 Voltage Offset Correction: v_off_ini = 2948.676 uV, v_off_fin = 2945.480 uV 08:43:17 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 267.100 uVs 08:43:19 I2Y Average and RMS of Measurements VT = -264.716 +- 6.799 uVs I2Y = -24.354 +- 0.626 uTm^2 08:43:30 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 08:43:34 Integrated Voltage Measurement... 08:43:43 Voltage Offset Correction: v_off_ini = 2950.899 uV, v_off_fin = 2943.544 uV 08:43:43 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -128.100 uVs 08:43:43 Integrated Voltage Measurement... 08:43:52 Voltage Offset Correction: v_off_ini = 2945.183 uV, v_off_fin = 2945.511 uV 08:43:52 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 101.522 uVs 08:43:52 Integrated Voltage Measurement... 08:44:01 Voltage Offset Correction: v_off_ini = 2944.102 uV, v_off_fin = 2937.519 uV 08:44:01 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -115.378 uVs 08:44:01 Integrated Voltage Measurement... 08:44:10 Voltage Offset Correction: v_off_ini = 2937.161 uV, v_off_fin = 2953.098 uV 08:44:10 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 119.892 uVs 08:44:10 Integrated Voltage Measurement... 08:44:19 Voltage Offset Correction: v_off_ini = 2951.945 uV, v_off_fin = 2931.707 uV 08:44:19 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -104.293 uVs 08:44:19 Integrated Voltage Measurement... 08:44:28 Voltage Offset Correction: v_off_ini = 2973.149 uV, v_off_fin = 2970.671 uV 08:44:28 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 104.062 uVs 08:44:33 I1X Average and RMS of Measurements VT = -112.208 +- 5.794 uVs I1X = 22.442 +- 1.159 uTm 08:44:35 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 08:44:39 Integrated Voltage Measurement... 08:44:48 Voltage Offset Correction: v_off_ini = 2953.813 uV, v_off_fin = 2940.638 uV 08:44:48 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -233.677 uVs 08:44:48 Integrated Voltage Measurement... 08:44:57 Voltage Offset Correction: v_off_ini = 2959.299 uV, v_off_fin = 2985.400 uV 08:44:57 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 246.827 uVs 08:44:57 Integrated Voltage Measurement... 08:45:06 Voltage Offset Correction: v_off_ini = 2975.170 uV, v_off_fin = 2957.001 uV 08:45:06 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -233.151 uVs 08:45:06 Integrated Voltage Measurement... 08:45:15 Voltage Offset Correction: v_off_ini = 2992.360 uV, v_off_fin = 2993.680 uV 08:45:15 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 232.697 uVs 08:45:15 Integrated Voltage Measurement... 08:45:24 Voltage Offset Correction: v_off_ini = 2983.269 uV, v_off_fin = 2965.505 uV 08:45:24 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -258.019 uVs 08:45:24 Integrated Voltage Measurement... 08:45:33 Voltage Offset Correction: v_off_ini = 2977.834 uV, v_off_fin = 2966.922 uV 08:45:33 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 239.300 uVs 08:45:38 I1Y Average and RMS of Measurements VT = -240.612 +- 6.429 uVs I1Y = -48.121 +- 1.284 uTm 08:45:43 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 08:45:45 Integrated Voltage Measurement... 08:45:54 Voltage Offset Correction: v_off_ini = 2975.671 uV, v_off_fin = 2976.605 uV 08:45:54 Integrated Voltage Measurement dXYus = -1.0001 mm, dXYds = 0.0000 mm dXY = -1.0001 mm VT = -98.095 uVs 08:45:54 Integrated Voltage Measurement... 08:46:03 Voltage Offset Correction: v_off_ini = 2976.421 uV, v_off_fin = 2976.017 uV 08:46:03 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 92.195 uVs 08:46:03 Integrated Voltage Measurement... 08:46:12 Voltage Offset Correction: v_off_ini = 2977.130 uV, v_off_fin = 2971.210 uV 08:46:12 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -111.693 uVs 08:46:12 Integrated Voltage Measurement... 08:46:21 Voltage Offset Correction: v_off_ini = 2977.374 uV, v_off_fin = 2989.803 uV 08:46:21 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 102.892 uVs 08:46:21 Integrated Voltage Measurement... 08:46:30 Voltage Offset Correction: v_off_ini = 2973.449 uV, v_off_fin = 2969.604 uV 08:46:30 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -81.692 uVs 08:46:30 Integrated Voltage Measurement... 08:46:39 Voltage Offset Correction: v_off_ini = 2987.852 uV, v_off_fin = 2978.106 uV 08:46:39 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 126.627 uVs 08:46:42 I2X Average and RMS of Measurements VT = -102.199 +- 5.149 uVs I2X = 9.402 +- 0.474 uTm^2 08:46:47 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 08:46:50 Integrated Voltage Measurement... 08:46:59 Voltage Offset Correction: v_off_ini = 2956.808 uV, v_off_fin = 2971.674 uV 08:46:59 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -314.525 uVs 08:46:59 Integrated Voltage Measurement... 08:47:08 Voltage Offset Correction: v_off_ini = 2974.821 uV, v_off_fin = 2976.220 uV 08:47:08 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 275.546 uVs 08:47:08 Integrated Voltage Measurement... 08:47:17 Voltage Offset Correction: v_off_ini = 3003.514 uV, v_off_fin = 2998.815 uV 08:47:17 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -293.655 uVs 08:47:17 Integrated Voltage Measurement... 08:47:26 Voltage Offset Correction: v_off_ini = 3010.627 uV, v_off_fin = 2996.797 uV 08:47:26 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 261.124 uVs 08:47:26 Integrated Voltage Measurement... 08:47:35 Voltage Offset Correction: v_off_ini = 2987.410 uV, v_off_fin = 2985.398 uV 08:47:35 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -284.958 uVs 08:47:35 Integrated Voltage Measurement... 08:47:44 Voltage Offset Correction: v_off_ini = 2990.604 uV, v_off_fin = 2991.387 uV 08:47:44 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 265.937 uVs 08:47:46 I2Y Average and RMS of Measurements VT = -282.624 +- 8.812 uVs I2Y = -26.001 +- 0.811 uTm^2 08:47:57 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 08:48:01 Integrated Voltage Measurement... 08:48:10 Voltage Offset Correction: v_off_ini = 3018.707 uV, v_off_fin = 3009.518 uV 08:48:10 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -86.136 uVs 08:48:10 Integrated Voltage Measurement... 08:48:19 Voltage Offset Correction: v_off_ini = 2971.249 uV, v_off_fin = 2984.032 uV 08:48:19 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 109.820 uVs 08:48:20 Integrated Voltage Measurement... 08:48:29 Voltage Offset Correction: v_off_ini = 2977.935 uV, v_off_fin = 2967.865 uV 08:48:29 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -100.621 uVs 08:48:29 Integrated Voltage Measurement... 08:48:38 Voltage Offset Correction: v_off_ini = 2983.847 uV, v_off_fin = 2976.580 uV 08:48:38 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 113.232 uVs 08:48:38 Integrated Voltage Measurement... 08:48:47 Voltage Offset Correction: v_off_ini = 2990.973 uV, v_off_fin = 2987.954 uV 08:48:47 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -92.555 uVs 08:48:47 Integrated Voltage Measurement... 08:48:56 Voltage Offset Correction: v_off_ini = 2998.303 uV, v_off_fin = 3000.344 uV 08:48:56 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 90.482 uVs 08:49:00 I1X Average and RMS of Measurements VT = -98.808 +- 6.318 uVs I1X = 19.762 +- 1.264 uTm 08:49:03 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 08:49:07 Integrated Voltage Measurement... 08:49:16 Voltage Offset Correction: v_off_ini = 2991.596 uV, v_off_fin = 2978.141 uV 08:49:16 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -249.425 uVs 08:49:16 Integrated Voltage Measurement... 08:49:25 Voltage Offset Correction: v_off_ini = 2984.379 uV, v_off_fin = 2994.409 uV 08:49:25 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 243.217 uVs 08:49:25 Integrated Voltage Measurement... 08:49:34 Voltage Offset Correction: v_off_ini = 2981.165 uV, v_off_fin = 2981.264 uV 08:49:34 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -243.425 uVs 08:49:34 Integrated Voltage Measurement... 08:49:43 Voltage Offset Correction: v_off_ini = 3001.447 uV, v_off_fin = 3010.142 uV 08:49:43 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 246.948 uVs 08:49:43 Integrated Voltage Measurement... 08:49:52 Voltage Offset Correction: v_off_ini = 2992.477 uV, v_off_fin = 2981.450 uV 08:49:52 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -235.950 uVs 08:49:52 Integrated Voltage Measurement... 08:50:01 Voltage Offset Correction: v_off_ini = 2999.423 uV, v_off_fin = 3005.901 uV 08:50:01 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 221.701 uVs 08:50:05 I1Y Average and RMS of Measurements VT = -240.111 +- 7.994 uVs I1Y = -48.021 +- 1.601 uTm 08:50:10 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 08:50:13 Integrated Voltage Measurement... 08:50:22 Voltage Offset Correction: v_off_ini = 2976.200 uV, v_off_fin = 2974.111 uV 08:50:22 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -124.690 uVs 08:50:22 Integrated Voltage Measurement... 08:50:31 Voltage Offset Correction: v_off_ini = 2983.968 uV, v_off_fin = 2997.043 uV 08:50:31 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 118.182 uVs 08:50:31 Integrated Voltage Measurement... 08:50:40 Voltage Offset Correction: v_off_ini = 2978.822 uV, v_off_fin = 2976.258 uV 08:50:40 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -116.050 uVs 08:50:40 Integrated Voltage Measurement... 08:50:49 Voltage Offset Correction: v_off_ini = 2978.022 uV, v_off_fin = 2993.685 uV 08:50:49 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 128.211 uVs 08:50:49 Integrated Voltage Measurement... 08:50:58 Voltage Offset Correction: v_off_ini = 2990.977 uV, v_off_fin = 2970.391 uV 08:50:58 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -129.392 uVs 08:50:58 Integrated Voltage Measurement... 08:51:07 Voltage Offset Correction: v_off_ini = 2976.357 uV, v_off_fin = 2986.188 uV 08:51:07 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 116.160 uVs 08:51:10 I2X Average and RMS of Measurements VT = -122.114 +- 0.547 uVs I2X = 11.234 +- 0.050 uTm^2 08:51:15 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 08:51:18 Integrated Voltage Measurement... 08:51:26 Voltage Offset Correction: v_off_ini = 2989.294 uV, v_off_fin = 2973.479 uV 08:51:27 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -272.551 uVs 08:51:27 Integrated Voltage Measurement... 08:51:36 Voltage Offset Correction: v_off_ini = 2988.542 uV, v_off_fin = 2998.128 uV 08:51:36 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 286.065 uVs 08:51:36 Integrated Voltage Measurement... 08:51:45 Voltage Offset Correction: v_off_ini = 3006.215 uV, v_off_fin = 3002.276 uV 08:51:45 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -286.354 uVs 08:51:45 Integrated Voltage Measurement... 08:51:54 Voltage Offset Correction: v_off_ini = 2994.346 uV, v_off_fin = 2990.162 uV 08:51:54 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 278.236 uVs 08:51:54 Integrated Voltage Measurement... 08:52:03 Voltage Offset Correction: v_off_ini = 2989.828 uV, v_off_fin = 2975.676 uV 08:52:03 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -295.775 uVs 08:52:03 Integrated Voltage Measurement... 08:52:12 Voltage Offset Correction: v_off_ini = 2984.213 uV, v_off_fin = 2977.650 uV 08:52:12 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 281.274 uVs 08:52:14 I2Y Average and RMS of Measurements VT = -283.376 +- 3.840 uVs I2Y = -26.071 +- 0.353 uTm^2 08:52:25 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 08:52:30 Integrated Voltage Measurement... 08:52:39 Voltage Offset Correction: v_off_ini = 2982.526 uV, v_off_fin = 3000.229 uV 08:52:39 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -107.288 uVs 08:52:39 Integrated Voltage Measurement... 08:52:48 Voltage Offset Correction: v_off_ini = 2987.374 uV, v_off_fin = 2963.868 uV 08:52:48 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 123.807 uVs 08:52:48 Integrated Voltage Measurement... 08:52:57 Voltage Offset Correction: v_off_ini = 2984.971 uV, v_off_fin = 2972.023 uV 08:52:57 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -143.103 uVs 08:52:57 Integrated Voltage Measurement... 08:53:06 Voltage Offset Correction: v_off_ini = 2992.672 uV, v_off_fin = 2990.932 uV 08:53:06 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 128.190 uVs 08:53:06 Integrated Voltage Measurement... 08:53:15 Voltage Offset Correction: v_off_ini = 2980.110 uV, v_off_fin = 2997.665 uV 08:53:15 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -116.901 uVs 08:53:15 Integrated Voltage Measurement... 08:53:24 Voltage Offset Correction: v_off_ini = 3000.834 uV, v_off_fin = 3007.668 uV 08:53:24 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 118.553 uVs 08:53:28 I1X Average and RMS of Measurements VT = -122.974 +- 9.005 uVs I1X = 24.595 +- 1.801 uTm 08:53:30 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 08:53:35 Integrated Voltage Measurement... 08:53:44 Voltage Offset Correction: v_off_ini = 2983.443 uV, v_off_fin = 2987.508 uV 08:53:44 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -228.270 uVs 08:53:44 Integrated Voltage Measurement... 08:53:53 Voltage Offset Correction: v_off_ini = 2986.947 uV, v_off_fin = 2992.069 uV 08:53:53 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 249.902 uVs 08:53:53 Integrated Voltage Measurement... 08:54:02 Voltage Offset Correction: v_off_ini = 2974.772 uV, v_off_fin = 2963.979 uV 08:54:02 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -239.260 uVs 08:54:02 Integrated Voltage Measurement... 08:54:11 Voltage Offset Correction: v_off_ini = 2980.745 uV, v_off_fin = 3002.342 uV 08:54:11 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 255.797 uVs 08:54:11 Integrated Voltage Measurement... 08:54:20 Voltage Offset Correction: v_off_ini = 2981.916 uV, v_off_fin = 2978.015 uV 08:54:20 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -243.420 uVs 08:54:20 Integrated Voltage Measurement... 08:54:29 Voltage Offset Correction: v_off_ini = 2976.449 uV, v_off_fin = 2985.453 uV 08:54:29 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 224.101 uVs 08:54:33 I1Y Average and RMS of Measurements VT = -240.125 +- 5.669 uVs I1Y = -48.020 +- 1.134 uTm 08:54:38 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 08:54:41 Integrated Voltage Measurement... 08:54:50 Voltage Offset Correction: v_off_ini = 2973.447 uV, v_off_fin = 2958.959 uV 08:54:50 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -142.248 uVs 08:54:50 Integrated Voltage Measurement... 08:54:59 Voltage Offset Correction: v_off_ini = 2981.753 uV, v_off_fin = 2982.179 uV 08:54:59 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 138.126 uVs 08:54:59 Integrated Voltage Measurement... 08:55:08 Voltage Offset Correction: v_off_ini = 2971.767 uV, v_off_fin = 2969.180 uV 08:55:08 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -137.628 uVs 08:55:08 Integrated Voltage Measurement... 08:55:17 Voltage Offset Correction: v_off_ini = 2987.240 uV, v_off_fin = 2992.143 uV 08:55:17 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 122.489 uVs 08:55:17 Integrated Voltage Measurement... 08:55:26 Voltage Offset Correction: v_off_ini = 2993.427 uV, v_off_fin = 2986.961 uV 08:55:26 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -134.967 uVs 08:55:26 Integrated Voltage Measurement... 08:55:35 Voltage Offset Correction: v_off_ini = 2982.694 uV, v_off_fin = 2994.490 uV 08:55:36 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 132.001 uVs 08:55:38 I2X Average and RMS of Measurements VT = -134.576 +- 4.207 uVs I2X = 12.381 +- 0.387 uTm^2 08:55:43 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 08:55:46 Integrated Voltage Measurement... 08:55:55 Voltage Offset Correction: v_off_ini = 2997.248 uV, v_off_fin = 3006.078 uV 08:55:55 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -290.712 uVs 08:55:55 Integrated Voltage Measurement... 08:56:04 Voltage Offset Correction: v_off_ini = 3004.082 uV, v_off_fin = 3018.658 uV 08:56:04 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 280.237 uVs 08:56:04 Integrated Voltage Measurement... 08:56:13 Voltage Offset Correction: v_off_ini = 3013.955 uV, v_off_fin = 3007.463 uV 08:56:13 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = -297.730 uVs 08:56:13 Integrated Voltage Measurement... 08:56:22 Voltage Offset Correction: v_off_ini = 2996.412 uV, v_off_fin = 3018.940 uV 08:56:22 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 298.614 uVs 08:56:22 Integrated Voltage Measurement... 08:56:31 Voltage Offset Correction: v_off_ini = 3028.829 uV, v_off_fin = 3014.303 uV 08:56:31 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -283.004 uVs 08:56:31 Integrated Voltage Measurement... 08:56:40 Voltage Offset Correction: v_off_ini = 3010.787 uV, v_off_fin = 3012.726 uV 08:56:40 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 326.164 uVs 08:56:42 I2Y Average and RMS of Measurements VT = -296.077 +- 7.941 uVs I2Y = -27.240 +- 0.731 uTm^2 08:56:53 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 08:56:57 Integrated Voltage Measurement... 08:57:06 Voltage Offset Correction: v_off_ini = 3022.355 uV, v_off_fin = 3013.077 uV 08:57:06 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.4999 mm dXY = -0.4999 mm VT = -140.469 uVs 08:57:06 Integrated Voltage Measurement... 08:57:15 Voltage Offset Correction: v_off_ini = 3011.640 uV, v_off_fin = 2994.847 uV 08:57:15 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 127.672 uVs 08:57:15 Integrated Voltage Measurement... 08:57:24 Voltage Offset Correction: v_off_ini = 3003.131 uV, v_off_fin = 2998.349 uV 08:57:24 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -145.708 uVs 08:57:24 Integrated Voltage Measurement... 08:57:33 Voltage Offset Correction: v_off_ini = 3003.333 uV, v_off_fin = 3001.751 uV 08:57:33 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 137.459 uVs 08:57:34 Integrated Voltage Measurement... 08:57:43 Voltage Offset Correction: v_off_ini = 3003.754 uV, v_off_fin = 3005.401 uV 08:57:43 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -115.163 uVs 08:57:43 Integrated Voltage Measurement... 08:57:52 Voltage Offset Correction: v_off_ini = 3006.187 uV, v_off_fin = 2998.023 uV 08:57:52 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 141.693 uVs 08:57:56 I1X Average and RMS of Measurements VT = -134.694 +- 5.389 uVs I1X = 26.940 +- 1.078 uTm 08:57:59 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 08:58:03 Integrated Voltage Measurement... 08:58:12 Voltage Offset Correction: v_off_ini = 2968.599 uV, v_off_fin = 2971.825 uV 08:58:12 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -241.184 uVs 08:58:12 Integrated Voltage Measurement... 08:58:21 Voltage Offset Correction: v_off_ini = 2995.936 uV, v_off_fin = 2993.690 uV 08:58:21 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 256.803 uVs 08:58:21 Integrated Voltage Measurement... 08:58:30 Voltage Offset Correction: v_off_ini = 2984.646 uV, v_off_fin = 2983.047 uV 08:58:30 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -263.242 uVs 08:58:30 Integrated Voltage Measurement... 08:58:39 Voltage Offset Correction: v_off_ini = 2990.970 uV, v_off_fin = 3019.505 uV 08:58:39 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 260.118 uVs 08:58:39 Integrated Voltage Measurement... 08:58:48 Voltage Offset Correction: v_off_ini = 3003.268 uV, v_off_fin = 2991.489 uV 08:58:48 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -253.941 uVs 08:58:48 Integrated Voltage Measurement... 08:58:57 Voltage Offset Correction: v_off_ini = 2998.470 uV, v_off_fin = 3028.193 uV 08:58:57 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 238.509 uVs 08:59:01 I1Y Average and RMS of Measurements VT = -252.300 +- 6.729 uVs I1Y = -50.458 +- 1.347 uTm 08:59:06 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 08:59:09 Integrated Voltage Measurement... 08:59:18 Voltage Offset Correction: v_off_ini = 2999.531 uV, v_off_fin = 2979.278 uV 08:59:18 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -112.579 uVs 08:59:18 Integrated Voltage Measurement... 08:59:27 Voltage Offset Correction: v_off_ini = 2990.644 uV, v_off_fin = 2979.753 uV 08:59:27 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 111.317 uVs 08:59:27 Integrated Voltage Measurement... 08:59:36 Voltage Offset Correction: v_off_ini = 2985.395 uV, v_off_fin = 2981.292 uV 08:59:36 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -123.832 uVs 08:59:36 Integrated Voltage Measurement... 08:59:45 Voltage Offset Correction: v_off_ini = 2990.443 uV, v_off_fin = 2989.599 uV 08:59:45 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 136.971 uVs 08:59:45 Integrated Voltage Measurement... 08:59:54 Voltage Offset Correction: v_off_ini = 2976.826 uV, v_off_fin = 2974.028 uV 08:59:54 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -145.662 uVs 08:59:54 Integrated Voltage Measurement... 09:00:03 Voltage Offset Correction: v_off_ini = 2991.966 uV, v_off_fin = 3008.474 uV 09:00:03 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 133.294 uVs 09:00:06 I2X Average and RMS of Measurements VT = -127.276 +- 11.454 uVs I2X = 11.709 +- 1.054 uTm^2 09:00:11 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:00:13 Integrated Voltage Measurement... 09:00:22 Voltage Offset Correction: v_off_ini = 3004.720 uV, v_off_fin = 3019.095 uV 09:00:22 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -288.681 uVs 09:00:23 Integrated Voltage Measurement... 09:00:31 Voltage Offset Correction: v_off_ini = 3015.273 uV, v_off_fin = 3028.040 uV 09:00:32 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 285.787 uVs 09:00:32 Integrated Voltage Measurement... 09:00:41 Voltage Offset Correction: v_off_ini = 3016.892 uV, v_off_fin = 3013.874 uV 09:00:41 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -293.648 uVs 09:00:41 Integrated Voltage Measurement... 09:00:50 Voltage Offset Correction: v_off_ini = 3022.539 uV, v_off_fin = 3025.953 uV 09:00:50 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 318.712 uVs 09:00:50 Integrated Voltage Measurement... 09:00:59 Voltage Offset Correction: v_off_ini = 3004.806 uV, v_off_fin = 2997.724 uV 09:00:59 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -303.938 uVs 09:00:59 Integrated Voltage Measurement... 09:01:08 Voltage Offset Correction: v_off_ini = 3001.762 uV, v_off_fin = 2994.215 uV 09:01:08 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 300.446 uVs 09:01:11 I2Y Average and RMS of Measurements VT = -298.535 +- 8.155 uVs I2Y = -27.465 +- 0.750 uTm^2 09:01:21 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:01:26 Integrated Voltage Measurement... 09:01:35 Voltage Offset Correction: v_off_ini = 3019.255 uV, v_off_fin = 3000.490 uV 09:01:35 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -154.636 uVs 09:01:35 Integrated Voltage Measurement... 09:01:44 Voltage Offset Correction: v_off_ini = 3008.136 uV, v_off_fin = 3005.454 uV 09:01:44 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 123.364 uVs 09:01:44 Integrated Voltage Measurement... 09:01:53 Voltage Offset Correction: v_off_ini = 3025.930 uV, v_off_fin = 3013.354 uV 09:01:53 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -140.185 uVs 09:01:53 Integrated Voltage Measurement... 09:02:02 Voltage Offset Correction: v_off_ini = 2999.941 uV, v_off_fin = 3002.968 uV 09:02:02 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 139.679 uVs 09:02:02 Integrated Voltage Measurement... 09:02:11 Voltage Offset Correction: v_off_ini = 3026.759 uV, v_off_fin = 3014.953 uV 09:02:11 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -155.663 uVs 09:02:11 Integrated Voltage Measurement... 09:02:20 Voltage Offset Correction: v_off_ini = 3020.258 uV, v_off_fin = 3002.445 uV 09:02:20 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 124.791 uVs 09:02:24 I1X Average and RMS of Measurements VT = -139.720 +- 0.523 uVs I1X = 27.944 +- 0.105 uTm 09:02:27 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:02:31 Integrated Voltage Measurement... 09:02:40 Voltage Offset Correction: v_off_ini = 3004.748 uV, v_off_fin = 2989.653 uV 09:02:40 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -247.783 uVs 09:02:40 Integrated Voltage Measurement... 09:02:49 Voltage Offset Correction: v_off_ini = 3005.290 uV, v_off_fin = 3013.516 uV 09:02:49 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 260.612 uVs 09:02:49 Integrated Voltage Measurement... 09:02:58 Voltage Offset Correction: v_off_ini = 3020.030 uV, v_off_fin = 3004.468 uV 09:02:58 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -244.177 uVs 09:02:58 Integrated Voltage Measurement... 09:03:07 Voltage Offset Correction: v_off_ini = 3011.277 uV, v_off_fin = 3029.452 uV 09:03:07 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 254.749 uVs 09:03:07 Integrated Voltage Measurement... 09:03:16 Voltage Offset Correction: v_off_ini = 3033.844 uV, v_off_fin = 3013.080 uV 09:03:16 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -264.750 uVs 09:03:16 Integrated Voltage Measurement... 09:03:25 Voltage Offset Correction: v_off_ini = 3010.649 uV, v_off_fin = 3025.056 uV 09:03:25 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 254.766 uVs 09:03:29 I1Y Average and RMS of Measurements VT = -254.473 +- 4.207 uVs I1Y = -50.890 +- 0.841 uTm 09:03:34 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:03:37 Integrated Voltage Measurement... 09:03:46 Voltage Offset Correction: v_off_ini = 3029.015 uV, v_off_fin = 3017.337 uV 09:03:46 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -146.810 uVs 09:03:46 Integrated Voltage Measurement... 09:03:55 Voltage Offset Correction: v_off_ini = 3026.630 uV, v_off_fin = 3023.996 uV 09:03:55 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 122.104 uVs 09:03:55 Integrated Voltage Measurement... 09:04:04 Voltage Offset Correction: v_off_ini = 3002.154 uV, v_off_fin = 2993.630 uV 09:04:04 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -156.607 uVs 09:04:04 Integrated Voltage Measurement... 09:04:13 Voltage Offset Correction: v_off_ini = 3006.504 uV, v_off_fin = 3016.540 uV 09:04:13 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 160.669 uVs 09:04:13 Integrated Voltage Measurement... 09:04:22 Voltage Offset Correction: v_off_ini = 2998.013 uV, v_off_fin = 2976.485 uV 09:04:22 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -128.442 uVs 09:04:22 Integrated Voltage Measurement... 09:04:31 Voltage Offset Correction: v_off_ini = 2991.668 uV, v_off_fin = 3010.184 uV 09:04:31 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 133.613 uVs 09:04:34 I2X Average and RMS of Measurements VT = -141.374 +- 12.287 uVs I2X = 13.006 +- 1.130 uTm^2 09:04:39 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:04:41 Integrated Voltage Measurement... 09:04:50 Voltage Offset Correction: v_off_ini = 3014.101 uV, v_off_fin = 3012.602 uV 09:04:50 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -303.262 uVs 09:04:51 Integrated Voltage Measurement... 09:05:00 Voltage Offset Correction: v_off_ini = 3019.501 uV, v_off_fin = 3021.906 uV 09:05:00 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 309.903 uVs 09:05:00 Integrated Voltage Measurement... 09:05:09 Voltage Offset Correction: v_off_ini = 3019.450 uV, v_off_fin = 3014.355 uV 09:05:09 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -329.186 uVs 09:05:09 Integrated Voltage Measurement... 09:05:18 Voltage Offset Correction: v_off_ini = 3007.211 uV, v_off_fin = 3018.279 uV 09:05:18 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 294.368 uVs 09:05:18 Integrated Voltage Measurement... 09:05:27 Voltage Offset Correction: v_off_ini = 3030.968 uV, v_off_fin = 3036.848 uV 09:05:27 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -297.082 uVs 09:05:27 Integrated Voltage Measurement... 09:05:36 Voltage Offset Correction: v_off_ini = 3040.235 uV, v_off_fin = 3045.133 uV 09:05:36 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 291.366 uVs 09:05:38 I2Y Average and RMS of Measurements VT = -304.195 +- 7.362 uVs I2Y = -27.986 +- 0.677 uTm^2 09:05:49 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:05:53 Integrated Voltage Measurement... 09:06:02 Voltage Offset Correction: v_off_ini = 3027.979 uV, v_off_fin = 3050.621 uV 09:06:02 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -153.209 uVs 09:06:02 Integrated Voltage Measurement... 09:06:11 Voltage Offset Correction: v_off_ini = 3033.984 uV, v_off_fin = 3034.087 uV 09:06:11 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 160.294 uVs 09:06:11 Integrated Voltage Measurement... 09:06:20 Voltage Offset Correction: v_off_ini = 3021.755 uV, v_off_fin = 3023.843 uV 09:06:20 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -159.731 uVs 09:06:20 Integrated Voltage Measurement... 09:06:29 Voltage Offset Correction: v_off_ini = 3023.921 uV, v_off_fin = 3016.683 uV 09:06:29 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 145.384 uVs 09:06:29 Integrated Voltage Measurement... 09:06:38 Voltage Offset Correction: v_off_ini = 3005.454 uV, v_off_fin = 3012.587 uV 09:06:38 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -149.129 uVs 09:06:38 Integrated Voltage Measurement... 09:06:47 Voltage Offset Correction: v_off_ini = 3003.383 uV, v_off_fin = 3004.522 uV 09:06:47 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 177.754 uVs 09:06:51 I1X Average and RMS of Measurements VT = -157.584 +- 4.482 uVs I1X = 31.517 +- 0.896 uTm 09:06:54 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:06:58 Integrated Voltage Measurement... 09:07:07 Voltage Offset Correction: v_off_ini = 3000.007 uV, v_off_fin = 2985.961 uV 09:07:07 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -247.335 uVs 09:07:07 Integrated Voltage Measurement... 09:07:16 Voltage Offset Correction: v_off_ini = 3036.987 uV, v_off_fin = 3055.970 uV 09:07:16 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 253.361 uVs 09:07:17 Integrated Voltage Measurement... 09:07:26 Voltage Offset Correction: v_off_ini = 3022.879 uV, v_off_fin = 2998.200 uV 09:07:26 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -256.240 uVs 09:07:26 Integrated Voltage Measurement... 09:07:35 Voltage Offset Correction: v_off_ini = 2999.685 uV, v_off_fin = 3019.580 uV 09:07:35 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = 253.433 uVs 09:07:35 Integrated Voltage Measurement... 09:07:44 Voltage Offset Correction: v_off_ini = 3009.250 uV, v_off_fin = 2997.171 uV 09:07:44 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -248.735 uVs 09:07:44 Integrated Voltage Measurement... 09:07:53 Voltage Offset Correction: v_off_ini = 3010.755 uV, v_off_fin = 3036.698 uV 09:07:53 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = 259.372 uVs 09:07:57 I1Y Average and RMS of Measurements VT = -253.079 +- 1.958 uVs I1Y = -50.618 +- 0.393 uTm 09:08:02 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:08:05 Integrated Voltage Measurement... 09:08:14 Voltage Offset Correction: v_off_ini = 3011.448 uV, v_off_fin = 3028.311 uV 09:08:14 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -150.306 uVs 09:08:14 Integrated Voltage Measurement... 09:08:23 Voltage Offset Correction: v_off_ini = 3037.658 uV, v_off_fin = 3036.695 uV 09:08:23 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 164.798 uVs 09:08:23 Integrated Voltage Measurement... 09:08:32 Voltage Offset Correction: v_off_ini = 3028.793 uV, v_off_fin = 3020.554 uV 09:08:32 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -170.404 uVs 09:08:32 Integrated Voltage Measurement... 09:08:41 Voltage Offset Correction: v_off_ini = 3018.614 uV, v_off_fin = 3030.708 uV 09:08:41 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 169.010 uVs 09:08:41 Integrated Voltage Measurement... 09:08:50 Voltage Offset Correction: v_off_ini = 3024.838 uV, v_off_fin = 3000.550 uV 09:08:50 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -154.715 uVs 09:08:50 Integrated Voltage Measurement... 09:08:59 Voltage Offset Correction: v_off_ini = 3017.528 uV, v_off_fin = 3036.968 uV 09:08:59 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 177.393 uVs 09:09:02 I2X Average and RMS of Measurements VT = -164.438 +- 5.092 uVs I2X = 15.128 +- 0.468 uTm^2 09:09:07 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:09:09 Integrated Voltage Measurement... 09:09:18 Voltage Offset Correction: v_off_ini = 3025.254 uV, v_off_fin = 3033.607 uV 09:09:18 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -304.091 uVs 09:09:18 Integrated Voltage Measurement... 09:09:27 Voltage Offset Correction: v_off_ini = 3050.246 uV, v_off_fin = 3026.345 uV 09:09:27 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 308.294 uVs 09:09:27 Integrated Voltage Measurement... 09:09:36 Voltage Offset Correction: v_off_ini = 3038.383 uV, v_off_fin = 3051.768 uV 09:09:36 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -299.252 uVs 09:09:36 Integrated Voltage Measurement... 09:09:45 Voltage Offset Correction: v_off_ini = 3048.956 uV, v_off_fin = 3042.600 uV 09:09:45 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 336.196 uVs 09:09:45 Integrated Voltage Measurement... 09:09:54 Voltage Offset Correction: v_off_ini = 3036.578 uV, v_off_fin = 3033.883 uV 09:09:55 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -299.332 uVs 09:09:55 Integrated Voltage Measurement... 09:10:04 Voltage Offset Correction: v_off_ini = 3036.619 uV, v_off_fin = 3022.769 uV 09:10:04 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 316.839 uVs 09:10:06 I2Y Average and RMS of Measurements VT = -310.667 +- 5.049 uVs I2Y = -28.581 +- 0.465 uTm^2 09:10:17 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:10:21 Integrated Voltage Measurement... 09:10:30 Voltage Offset Correction: v_off_ini = 3004.754 uV, v_off_fin = 3001.803 uV 09:10:30 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -148.620 uVs 09:10:30 Integrated Voltage Measurement... 09:10:39 Voltage Offset Correction: v_off_ini = 3001.667 uV, v_off_fin = 3015.469 uV 09:10:39 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 166.408 uVs 09:10:39 Integrated Voltage Measurement... 09:10:48 Voltage Offset Correction: v_off_ini = 3024.719 uV, v_off_fin = 3014.561 uV 09:10:48 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -156.980 uVs 09:10:48 Integrated Voltage Measurement... 09:10:57 Voltage Offset Correction: v_off_ini = 3005.138 uV, v_off_fin = 3019.005 uV 09:10:57 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 152.264 uVs 09:10:57 Integrated Voltage Measurement... 09:11:06 Voltage Offset Correction: v_off_ini = 3018.654 uV, v_off_fin = 3015.388 uV 09:11:06 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -163.824 uVs 09:11:06 Integrated Voltage Measurement... 09:11:15 Voltage Offset Correction: v_off_ini = 3034.835 uV, v_off_fin = 3026.749 uV 09:11:15 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 161.324 uVs 09:11:20 I1X Average and RMS of Measurements VT = -158.237 +- 3.286 uVs I1X = 31.647 +- 0.657 uTm 09:11:22 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:11:26 Integrated Voltage Measurement... 09:11:35 Voltage Offset Correction: v_off_ini = 3021.235 uV, v_off_fin = 3002.377 uV 09:11:35 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -258.307 uVs 09:11:35 Integrated Voltage Measurement... 09:11:44 Voltage Offset Correction: v_off_ini = 3009.914 uV, v_off_fin = 3012.099 uV 09:11:44 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 253.396 uVs 09:11:44 Integrated Voltage Measurement... 09:11:53 Voltage Offset Correction: v_off_ini = 3014.121 uV, v_off_fin = 3010.802 uV 09:11:53 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -260.116 uVs 09:11:53 Integrated Voltage Measurement... 09:12:02 Voltage Offset Correction: v_off_ini = 3018.649 uV, v_off_fin = 3016.891 uV 09:12:02 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 252.450 uVs 09:12:02 Integrated Voltage Measurement... 09:12:11 Voltage Offset Correction: v_off_ini = 3032.455 uV, v_off_fin = 2995.483 uV 09:12:11 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -251.366 uVs 09:12:11 Integrated Voltage Measurement... 09:12:20 Voltage Offset Correction: v_off_ini = 3003.919 uV, v_off_fin = 2997.559 uV 09:12:20 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 280.729 uVs 09:12:24 I1Y Average and RMS of Measurements VT = -259.394 +- 4.708 uVs I1Y = -51.878 +- 0.942 uTm 09:12:29 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:12:32 Integrated Voltage Measurement... 09:12:41 Voltage Offset Correction: v_off_ini = 3018.856 uV, v_off_fin = 3014.424 uV 09:12:41 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -168.446 uVs 09:12:41 Integrated Voltage Measurement... 09:12:50 Voltage Offset Correction: v_off_ini = 3027.167 uV, v_off_fin = 3035.885 uV 09:12:50 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 155.529 uVs 09:12:50 Integrated Voltage Measurement... 09:12:59 Voltage Offset Correction: v_off_ini = 3011.882 uV, v_off_fin = 2997.538 uV 09:12:59 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -179.088 uVs 09:12:59 Integrated Voltage Measurement... 09:13:08 Voltage Offset Correction: v_off_ini = 3025.632 uV, v_off_fin = 3028.758 uV 09:13:09 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 168.226 uVs 09:13:09 Integrated Voltage Measurement... 09:13:18 Voltage Offset Correction: v_off_ini = 3031.275 uV, v_off_fin = 3014.222 uV 09:13:18 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -193.633 uVs 09:13:18 Integrated Voltage Measurement... 09:13:27 Voltage Offset Correction: v_off_ini = 3024.840 uV, v_off_fin = 3044.163 uV 09:13:27 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 143.377 uVs 09:13:29 I2X Average and RMS of Measurements VT = -168.050 +- 4.775 uVs I2X = 15.461 +- 0.439 uTm^2 09:13:34 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:13:37 Integrated Voltage Measurement... 09:13:46 Voltage Offset Correction: v_off_ini = 3041.899 uV, v_off_fin = 3035.466 uV 09:13:46 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -306.528 uVs 09:13:46 Integrated Voltage Measurement... 09:13:55 Voltage Offset Correction: v_off_ini = 3028.492 uV, v_off_fin = 3034.822 uV 09:13:55 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 302.425 uVs 09:13:55 Integrated Voltage Measurement... 09:14:04 Voltage Offset Correction: v_off_ini = 3002.679 uV, v_off_fin = 3017.087 uV 09:14:04 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -319.805 uVs 09:14:04 Integrated Voltage Measurement... 09:14:13 Voltage Offset Correction: v_off_ini = 3019.962 uV, v_off_fin = 3024.133 uV 09:14:13 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 329.496 uVs 09:14:13 Integrated Voltage Measurement... 09:14:22 Voltage Offset Correction: v_off_ini = 3025.098 uV, v_off_fin = 3020.198 uV 09:14:22 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -323.723 uVs 09:14:22 Integrated Voltage Measurement... 09:14:31 Voltage Offset Correction: v_off_ini = 3018.740 uV, v_off_fin = 3022.704 uV 09:14:32 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 309.388 uVs 09:14:34 I2Y Average and RMS of Measurements VT = -315.228 +- 8.289 uVs I2Y = -29.001 +- 0.763 uTm^2 09:14:46 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:14:50 Integrated Voltage Measurement... 09:14:59 Voltage Offset Correction: v_off_ini = 3049.209 uV, v_off_fin = 3039.486 uV 09:14:59 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -200.260 uVs 09:14:59 Integrated Voltage Measurement... 09:15:08 Voltage Offset Correction: v_off_ini = 3039.601 uV, v_off_fin = 3031.672 uV 09:15:08 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 174.262 uVs 09:15:09 Integrated Voltage Measurement... 09:15:18 Voltage Offset Correction: v_off_ini = 3025.345 uV, v_off_fin = 3034.266 uV 09:15:18 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -174.727 uVs 09:15:18 Integrated Voltage Measurement... 09:15:27 Voltage Offset Correction: v_off_ini = 3046.964 uV, v_off_fin = 3042.372 uV 09:15:27 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 163.608 uVs 09:15:27 Integrated Voltage Measurement... 09:15:36 Voltage Offset Correction: v_off_ini = 3030.281 uV, v_off_fin = 3040.004 uV 09:15:36 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -165.646 uVs 09:15:36 Integrated Voltage Measurement... 09:15:45 Voltage Offset Correction: v_off_ini = 3024.120 uV, v_off_fin = 3030.083 uV 09:15:45 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 171.557 uVs 09:15:49 I1X Average and RMS of Measurements VT = -175.010 +- 8.666 uVs I1X = 35.002 +- 1.733 uTm 09:15:52 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:15:56 Integrated Voltage Measurement... 09:16:05 Voltage Offset Correction: v_off_ini = 3021.655 uV, v_off_fin = 3000.707 uV 09:16:05 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -274.366 uVs 09:16:05 Integrated Voltage Measurement... 09:16:14 Voltage Offset Correction: v_off_ini = 3004.942 uV, v_off_fin = 3001.275 uV 09:16:14 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 273.817 uVs 09:16:14 Integrated Voltage Measurement... 09:16:23 Voltage Offset Correction: v_off_ini = 3003.037 uV, v_off_fin = 3001.035 uV 09:16:23 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -253.071 uVs 09:16:23 Integrated Voltage Measurement... 09:16:32 Voltage Offset Correction: v_off_ini = 3003.371 uV, v_off_fin = 3013.571 uV 09:16:32 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 278.485 uVs 09:16:32 Integrated Voltage Measurement... 09:16:41 Voltage Offset Correction: v_off_ini = 3013.524 uV, v_off_fin = 2983.856 uV 09:16:41 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -283.027 uVs 09:16:41 Integrated Voltage Measurement... 09:16:50 Voltage Offset Correction: v_off_ini = 3020.969 uV, v_off_fin = 3027.706 uV 09:16:50 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 278.770 uVs 09:16:54 I1Y Average and RMS of Measurements VT = -273.589 +- 6.183 uVs I1Y = -54.715 +- 1.239 uTm 09:16:59 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:17:02 Integrated Voltage Measurement... 09:17:11 Voltage Offset Correction: v_off_ini = 3006.232 uV, v_off_fin = 3005.470 uV 09:17:11 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -219.436 uVs 09:17:11 Integrated Voltage Measurement... 09:17:20 Voltage Offset Correction: v_off_ini = 3032.282 uV, v_off_fin = 3024.968 uV 09:17:20 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 181.836 uVs 09:17:20 Integrated Voltage Measurement... 09:17:29 Voltage Offset Correction: v_off_ini = 2998.067 uV, v_off_fin = 2993.524 uV 09:17:29 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -162.194 uVs 09:17:29 Integrated Voltage Measurement... 09:17:38 Voltage Offset Correction: v_off_ini = 3016.078 uV, v_off_fin = 3022.082 uV 09:17:38 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 168.293 uVs 09:17:38 Integrated Voltage Measurement... 09:17:47 Voltage Offset Correction: v_off_ini = 3021.424 uV, v_off_fin = 3001.929 uV 09:17:47 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -194.049 uVs 09:17:47 Integrated Voltage Measurement... 09:17:56 Voltage Offset Correction: v_off_ini = 3034.381 uV, v_off_fin = 3042.616 uV 09:17:56 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 159.292 uVs 09:17:59 I2X Average and RMS of Measurements VT = -180.850 +- 14.748 uVs I2X = 16.638 +- 1.357 uTm^2 09:18:04 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:18:06 Integrated Voltage Measurement... 09:18:15 Voltage Offset Correction: v_off_ini = 3021.456 uV, v_off_fin = 3019.486 uV 09:18:15 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -313.862 uVs 09:18:16 Integrated Voltage Measurement... 09:18:25 Voltage Offset Correction: v_off_ini = 3021.039 uV, v_off_fin = 3022.743 uV 09:18:25 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 328.576 uVs 09:18:25 Integrated Voltage Measurement... 09:18:34 Voltage Offset Correction: v_off_ini = 3023.536 uV, v_off_fin = 3006.304 uV 09:18:34 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -341.015 uVs 09:18:34 Integrated Voltage Measurement... 09:18:43 Voltage Offset Correction: v_off_ini = 3020.809 uV, v_off_fin = 3033.116 uV 09:18:43 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 332.944 uVs 09:18:43 Integrated Voltage Measurement... 09:18:52 Voltage Offset Correction: v_off_ini = 3017.694 uV, v_off_fin = 2998.839 uV 09:18:52 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -318.109 uVs 09:18:52 Integrated Voltage Measurement... 09:19:01 Voltage Offset Correction: v_off_ini = 3010.417 uV, v_off_fin = 3006.802 uV 09:19:01 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 325.969 uVs 09:19:04 I2Y Average and RMS of Measurements VT = -326.746 +- 7.244 uVs I2Y = -30.061 +- 0.666 uTm^2 09:19:16 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:19:20 Integrated Voltage Measurement... 09:19:29 Voltage Offset Correction: v_off_ini = 3010.967 uV, v_off_fin = 3012.175 uV 09:19:29 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -189.285 uVs 09:19:29 Integrated Voltage Measurement... 09:19:38 Voltage Offset Correction: v_off_ini = 3010.643 uV, v_off_fin = 2999.919 uV 09:19:38 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 174.621 uVs 09:19:38 Integrated Voltage Measurement... 09:19:47 Voltage Offset Correction: v_off_ini = 2997.445 uV, v_off_fin = 2972.245 uV 09:19:47 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -197.248 uVs 09:19:47 Integrated Voltage Measurement... 09:19:56 Voltage Offset Correction: v_off_ini = 2991.207 uV, v_off_fin = 2999.968 uV 09:19:56 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 191.381 uVs 09:19:56 Integrated Voltage Measurement... 09:20:05 Voltage Offset Correction: v_off_ini = 2984.461 uV, v_off_fin = 2985.594 uV 09:20:05 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -186.064 uVs 09:20:05 Integrated Voltage Measurement... 09:20:14 Voltage Offset Correction: v_off_ini = 2982.879 uV, v_off_fin = 2972.404 uV 09:20:14 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 178.352 uVs 09:20:18 I1X Average and RMS of Measurements VT = -186.158 +- 5.768 uVs I1X = 37.232 +- 1.154 uTm 09:20:21 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:20:25 Integrated Voltage Measurement... 09:20:34 Voltage Offset Correction: v_off_ini = 2982.796 uV, v_off_fin = 2965.421 uV 09:20:34 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = -274.957 uVs 09:20:34 Integrated Voltage Measurement... 09:20:43 Voltage Offset Correction: v_off_ini = 2975.076 uV, v_off_fin = 2993.660 uV 09:20:43 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 277.625 uVs 09:20:43 Integrated Voltage Measurement... 09:20:52 Voltage Offset Correction: v_off_ini = 2984.855 uV, v_off_fin = 2958.049 uV 09:20:52 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -263.428 uVs 09:20:52 Integrated Voltage Measurement... 09:21:01 Voltage Offset Correction: v_off_ini = 2959.014 uV, v_off_fin = 2976.990 uV 09:21:01 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 259.829 uVs 09:21:01 Integrated Voltage Measurement... 09:21:10 Voltage Offset Correction: v_off_ini = 2974.166 uV, v_off_fin = 2958.025 uV 09:21:10 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -284.649 uVs 09:21:10 Integrated Voltage Measurement... 09:21:19 Voltage Offset Correction: v_off_ini = 2967.014 uV, v_off_fin = 2996.435 uV 09:21:19 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 290.699 uVs 09:21:23 I1Y Average and RMS of Measurements VT = -275.198 +- 10.661 uVs I1Y = -55.040 +- 2.132 uTm 09:21:28 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:21:31 Integrated Voltage Measurement... 09:21:40 Voltage Offset Correction: v_off_ini = 2971.108 uV, v_off_fin = 2953.037 uV 09:21:40 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -178.404 uVs 09:21:40 Integrated Voltage Measurement... 09:21:49 Voltage Offset Correction: v_off_ini = 2950.215 uV, v_off_fin = 2971.063 uV 09:21:49 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 201.052 uVs 09:21:49 Integrated Voltage Measurement... 09:21:58 Voltage Offset Correction: v_off_ini = 2954.887 uV, v_off_fin = 2961.610 uV 09:21:58 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -203.251 uVs 09:21:58 Integrated Voltage Measurement... 09:22:07 Voltage Offset Correction: v_off_ini = 2932.860 uV, v_off_fin = 2957.350 uV 09:22:07 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 192.588 uVs 09:22:07 Integrated Voltage Measurement... 09:22:16 Voltage Offset Correction: v_off_ini = 2952.386 uV, v_off_fin = 2934.413 uV 09:22:16 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -198.310 uVs 09:22:16 Integrated Voltage Measurement... 09:22:25 Voltage Offset Correction: v_off_ini = 2954.886 uV, v_off_fin = 2960.248 uV 09:22:25 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 178.657 uVs 09:22:27 I2X Average and RMS of Measurements VT = -192.044 +- 4.186 uVs I2X = 17.668 +- 0.385 uTm^2 09:22:32 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:22:35 Integrated Voltage Measurement... 09:22:44 Voltage Offset Correction: v_off_ini = 2965.668 uV, v_off_fin = 2989.528 uV 09:22:44 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -354.342 uVs 09:22:44 Integrated Voltage Measurement... 09:22:53 Voltage Offset Correction: v_off_ini = 2985.015 uV, v_off_fin = 2975.948 uV 09:22:53 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 317.704 uVs 09:22:53 Integrated Voltage Measurement... 09:23:02 Voltage Offset Correction: v_off_ini = 2969.498 uV, v_off_fin = 2973.083 uV 09:23:02 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -313.433 uVs 09:23:02 Integrated Voltage Measurement... 09:23:11 Voltage Offset Correction: v_off_ini = 2968.371 uV, v_off_fin = 2964.960 uV 09:23:11 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 299.600 uVs 09:23:11 Integrated Voltage Measurement... 09:23:20 Voltage Offset Correction: v_off_ini = 2963.156 uV, v_off_fin = 2978.739 uV 09:23:20 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -325.810 uVs 09:23:20 Integrated Voltage Measurement... 09:23:29 Voltage Offset Correction: v_off_ini = 2965.597 uV, v_off_fin = 2965.717 uV 09:23:29 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 326.111 uVs 09:23:32 I2Y Average and RMS of Measurements VT = -322.833 +- 12.247 uVs I2Y = -29.701 +- 1.127 uTm^2 09:23:46 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:23:50 Integrated Voltage Measurement... 09:23:59 Voltage Offset Correction: v_off_ini = 2941.184 uV, v_off_fin = 2955.957 uV 09:23:59 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -208.179 uVs 09:23:59 Integrated Voltage Measurement... 09:24:08 Voltage Offset Correction: v_off_ini = 2978.662 uV, v_off_fin = 2979.888 uV 09:24:08 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 201.038 uVs 09:24:09 Integrated Voltage Measurement... 09:24:18 Voltage Offset Correction: v_off_ini = 2967.933 uV, v_off_fin = 2979.995 uV 09:24:18 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -190.312 uVs 09:24:18 Integrated Voltage Measurement... 09:24:27 Voltage Offset Correction: v_off_ini = 2969.335 uV, v_off_fin = 2974.514 uV 09:24:27 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 201.713 uVs 09:24:27 Integrated Voltage Measurement... 09:24:36 Voltage Offset Correction: v_off_ini = 2971.812 uV, v_off_fin = 2969.579 uV 09:24:36 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -205.672 uVs 09:24:36 Integrated Voltage Measurement... 09:24:45 Voltage Offset Correction: v_off_ini = 2970.082 uV, v_off_fin = 2976.806 uV 09:24:45 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 160.724 uVs 09:24:49 I1X Average and RMS of Measurements VT = -194.606 +- 8.797 uVs I1X = 38.921 +- 1.759 uTm 09:24:51 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:24:56 Integrated Voltage Measurement... 09:25:05 Voltage Offset Correction: v_off_ini = 2964.162 uV, v_off_fin = 2963.542 uV 09:25:05 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -253.115 uVs 09:25:05 Integrated Voltage Measurement... 09:25:14 Voltage Offset Correction: v_off_ini = 2968.655 uV, v_off_fin = 2982.908 uV 09:25:14 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 270.640 uVs 09:25:14 Integrated Voltage Measurement... 09:25:23 Voltage Offset Correction: v_off_ini = 2963.738 uV, v_off_fin = 2962.342 uV 09:25:23 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -276.080 uVs 09:25:23 Integrated Voltage Measurement... 09:25:32 Voltage Offset Correction: v_off_ini = 2965.169 uV, v_off_fin = 2986.212 uV 09:25:32 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 277.690 uVs 09:25:32 Integrated Voltage Measurement... 09:25:41 Voltage Offset Correction: v_off_ini = 2963.913 uV, v_off_fin = 2952.454 uV 09:25:41 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -262.939 uVs 09:25:41 Integrated Voltage Measurement... 09:25:50 Voltage Offset Correction: v_off_ini = 2957.040 uV, v_off_fin = 3001.487 uV 09:25:50 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 276.872 uVs 09:25:54 I1Y Average and RMS of Measurements VT = -269.556 +- 6.132 uVs I1Y = -53.910 +- 1.228 uTm 09:25:59 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:26:01 Integrated Voltage Measurement... 09:26:10 Voltage Offset Correction: v_off_ini = 2979.902 uV, v_off_fin = 2971.424 uV 09:26:10 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -165.714 uVs 09:26:11 Integrated Voltage Measurement... 09:26:19 Voltage Offset Correction: v_off_ini = 2969.628 uV, v_off_fin = 2990.040 uV 09:26:19 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 202.172 uVs 09:26:20 Integrated Voltage Measurement... 09:26:29 Voltage Offset Correction: v_off_ini = 2968.424 uV, v_off_fin = 2959.965 uV 09:26:29 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -192.374 uVs 09:26:29 Integrated Voltage Measurement... 09:26:38 Voltage Offset Correction: v_off_ini = 2961.838 uV, v_off_fin = 2992.808 uV 09:26:38 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 229.019 uVs 09:26:38 Integrated Voltage Measurement... 09:26:47 Voltage Offset Correction: v_off_ini = 2984.332 uV, v_off_fin = 2960.721 uV 09:26:47 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -197.360 uVs 09:26:47 Integrated Voltage Measurement... 09:26:56 Voltage Offset Correction: v_off_ini = 2953.263 uV, v_off_fin = 2972.590 uV 09:26:56 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 187.371 uVs 09:26:58 I2X Average and RMS of Measurements VT = -195.668 +- 11.169 uVs I2X = 18.002 +- 1.028 uTm^2 09:27:03 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:27:06 Integrated Voltage Measurement... 09:27:15 Voltage Offset Correction: v_off_ini = 2957.795 uV, v_off_fin = 2983.202 uV 09:27:15 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -305.634 uVs 09:27:15 Integrated Voltage Measurement... 09:27:24 Voltage Offset Correction: v_off_ini = 2979.273 uV, v_off_fin = 2964.040 uV 09:27:24 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 306.435 uVs 09:27:24 Integrated Voltage Measurement... 09:27:33 Voltage Offset Correction: v_off_ini = 2937.758 uV, v_off_fin = 2944.365 uV 09:27:33 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -333.696 uVs 09:27:33 Integrated Voltage Measurement... 09:27:42 Voltage Offset Correction: v_off_ini = 2947.786 uV, v_off_fin = 2936.371 uV 09:27:42 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 320.684 uVs 09:27:42 Integrated Voltage Measurement... 09:27:51 Voltage Offset Correction: v_off_ini = 2944.005 uV, v_off_fin = 2947.930 uV 09:27:51 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -306.426 uVs 09:27:52 Integrated Voltage Measurement... 09:28:01 Voltage Offset Correction: v_off_ini = 2929.208 uV, v_off_fin = 2938.349 uV 09:28:01 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 322.187 uVs 09:28:03 I2Y Average and RMS of Measurements VT = -315.844 +- 8.705 uVs I2Y = -29.058 +- 0.801 uTm^2 09:28:17 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:28:22 Integrated Voltage Measurement... 09:28:31 Voltage Offset Correction: v_off_ini = 2949.945 uV, v_off_fin = 2933.045 uV 09:28:31 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -207.694 uVs 09:28:31 Integrated Voltage Measurement... 09:28:40 Voltage Offset Correction: v_off_ini = 2919.071 uV, v_off_fin = 2905.205 uV 09:28:40 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 180.924 uVs 09:28:40 Integrated Voltage Measurement... 09:28:49 Voltage Offset Correction: v_off_ini = 2907.350 uV, v_off_fin = 2922.790 uV 09:28:49 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -166.762 uVs 09:28:49 Integrated Voltage Measurement... 09:28:58 Voltage Offset Correction: v_off_ini = 2922.630 uV, v_off_fin = 2918.061 uV 09:28:58 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 185.135 uVs 09:28:58 Integrated Voltage Measurement... 09:29:07 Voltage Offset Correction: v_off_ini = 2912.168 uV, v_off_fin = 2905.007 uV 09:29:07 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -183.561 uVs 09:29:07 Integrated Voltage Measurement... 09:29:16 Voltage Offset Correction: v_off_ini = 2950.866 uV, v_off_fin = 2927.861 uV 09:29:16 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 178.506 uVs 09:29:20 I1X Average and RMS of Measurements VT = -183.764 +- 7.740 uVs I1X = 36.753 +- 1.548 uTm 09:29:22 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:29:27 Integrated Voltage Measurement... 09:29:36 Voltage Offset Correction: v_off_ini = 2935.498 uV, v_off_fin = 2919.794 uV 09:29:36 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -269.378 uVs 09:29:36 Integrated Voltage Measurement... 09:29:45 Voltage Offset Correction: v_off_ini = 2933.698 uV, v_off_fin = 2954.699 uV 09:29:45 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 272.366 uVs 09:29:45 Integrated Voltage Measurement... 09:29:54 Voltage Offset Correction: v_off_ini = 2946.060 uV, v_off_fin = 2924.157 uV 09:29:54 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -259.000 uVs 09:29:54 Integrated Voltage Measurement... 09:30:02 Voltage Offset Correction: v_off_ini = 2929.616 uV, v_off_fin = 2949.907 uV 09:30:03 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 264.856 uVs 09:30:03 Integrated Voltage Measurement... 09:30:11 Voltage Offset Correction: v_off_ini = 2929.123 uV, v_off_fin = 2907.043 uV 09:30:12 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -263.675 uVs 09:30:12 Integrated Voltage Measurement... 09:30:20 Voltage Offset Correction: v_off_ini = 2920.620 uV, v_off_fin = 2958.150 uV 09:30:21 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 266.820 uVs 09:30:25 I1Y Average and RMS of Measurements VT = -266.016 +- 3.691 uVs I1Y = -53.199 +- 0.739 uTm 09:30:30 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:30:32 Integrated Voltage Measurement... 09:30:41 Voltage Offset Correction: v_off_ini = 2956.781 uV, v_off_fin = 2948.552 uV 09:30:41 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -184.244 uVs 09:30:41 Integrated Voltage Measurement... 09:30:50 Voltage Offset Correction: v_off_ini = 2950.955 uV, v_off_fin = 2968.619 uV 09:30:50 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 210.896 uVs 09:30:50 Integrated Voltage Measurement... 09:30:59 Voltage Offset Correction: v_off_ini = 2949.930 uV, v_off_fin = 2962.393 uV 09:30:59 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -200.875 uVs 09:31:00 Integrated Voltage Measurement... 09:31:08 Voltage Offset Correction: v_off_ini = 2970.637 uV, v_off_fin = 2974.320 uV 09:31:09 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 200.286 uVs 09:31:09 Integrated Voltage Measurement... 09:31:18 Voltage Offset Correction: v_off_ini = 2968.365 uV, v_off_fin = 2947.936 uV 09:31:18 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -198.285 uVs 09:31:18 Integrated Voltage Measurement... 09:31:27 Voltage Offset Correction: v_off_ini = 2972.724 uV, v_off_fin = 2964.960 uV 09:31:27 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 190.006 uVs 09:31:29 I2X Average and RMS of Measurements VT = -197.432 +- 2.629 uVs I2X = 18.164 +- 0.241 uTm^2 09:31:34 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:31:37 Integrated Voltage Measurement... 09:31:46 Voltage Offset Correction: v_off_ini = 2942.017 uV, v_off_fin = 2930.148 uV 09:31:46 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -315.013 uVs 09:31:46 Integrated Voltage Measurement... 09:31:55 Voltage Offset Correction: v_off_ini = 2955.475 uV, v_off_fin = 2961.851 uV 09:31:55 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 291.418 uVs 09:31:55 Integrated Voltage Measurement... 09:32:04 Voltage Offset Correction: v_off_ini = 2954.158 uV, v_off_fin = 2942.608 uV 09:32:04 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -315.868 uVs 09:32:04 Integrated Voltage Measurement... 09:32:13 Voltage Offset Correction: v_off_ini = 2954.230 uV, v_off_fin = 2969.198 uV 09:32:13 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = 311.416 uVs 09:32:13 Integrated Voltage Measurement... 09:32:22 Voltage Offset Correction: v_off_ini = 2957.294 uV, v_off_fin = 2957.134 uV 09:32:22 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -304.515 uVs 09:32:22 Integrated Voltage Measurement... 09:32:31 Voltage Offset Correction: v_off_ini = 2944.060 uV, v_off_fin = 2930.227 uV 09:32:31 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 340.524 uVs 09:32:34 I2Y Average and RMS of Measurements VT = -313.126 +- 7.889 uVs I2Y = -28.808 +- 0.726 uTm^2 09:32:48 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:32:52 Integrated Voltage Measurement... 09:33:01 Voltage Offset Correction: v_off_ini = 2939.478 uV, v_off_fin = 2940.845 uV 09:33:01 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = -170.692 uVs 09:33:01 Integrated Voltage Measurement... 09:33:10 Voltage Offset Correction: v_off_ini = 2928.204 uV, v_off_fin = 2944.275 uV 09:33:10 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 187.911 uVs 09:33:10 Integrated Voltage Measurement... 09:33:19 Voltage Offset Correction: v_off_ini = 2946.462 uV, v_off_fin = 2936.968 uV 09:33:19 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -171.775 uVs 09:33:19 Integrated Voltage Measurement... 09:33:28 Voltage Offset Correction: v_off_ini = 2942.033 uV, v_off_fin = 2938.004 uV 09:33:28 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 198.073 uVs 09:33:28 Integrated Voltage Measurement... 09:33:37 Voltage Offset Correction: v_off_ini = 2940.843 uV, v_off_fin = 2943.596 uV 09:33:37 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -183.599 uVs 09:33:37 Integrated Voltage Measurement... 09:33:46 Voltage Offset Correction: v_off_ini = 2938.300 uV, v_off_fin = 2931.032 uV 09:33:46 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 184.626 uVs 09:33:51 I1X Average and RMS of Measurements VT = -182.779 +- 2.481 uVs I1X = 36.556 +- 0.498 uTm 09:33:53 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:33:57 Integrated Voltage Measurement... 09:34:06 Voltage Offset Correction: v_off_ini = 2971.890 uV, v_off_fin = 2945.867 uV 09:34:06 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -270.202 uVs 09:34:06 Integrated Voltage Measurement... 09:34:15 Voltage Offset Correction: v_off_ini = 2979.304 uV, v_off_fin = 2979.488 uV 09:34:15 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 252.126 uVs 09:34:15 Integrated Voltage Measurement... 09:34:24 Voltage Offset Correction: v_off_ini = 2964.399 uV, v_off_fin = 2971.950 uV 09:34:24 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -266.996 uVs 09:34:24 Integrated Voltage Measurement... 09:34:33 Voltage Offset Correction: v_off_ini = 2948.046 uV, v_off_fin = 2952.831 uV 09:34:33 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 274.492 uVs 09:34:33 Integrated Voltage Measurement... 09:34:42 Voltage Offset Correction: v_off_ini = 2953.440 uV, v_off_fin = 2933.280 uV 09:34:42 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -260.353 uVs 09:34:42 Integrated Voltage Measurement... 09:34:51 Voltage Offset Correction: v_off_ini = 2936.976 uV, v_off_fin = 2957.676 uV 09:34:51 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 276.565 uVs 09:34:55 I1Y Average and RMS of Measurements VT = -266.789 +- 4.085 uVs I1Y = -53.353 +- 0.816 uTm 09:35:00 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:35:03 Integrated Voltage Measurement... 09:35:12 Voltage Offset Correction: v_off_ini = 2924.494 uV, v_off_fin = 2921.994 uV 09:35:12 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -189.084 uVs 09:35:12 Integrated Voltage Measurement... 09:35:21 Voltage Offset Correction: v_off_ini = 2938.552 uV, v_off_fin = 2925.687 uV 09:35:21 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 185.731 uVs 09:35:21 Integrated Voltage Measurement... 09:35:31 Voltage Offset Correction: v_off_ini = 2922.863 uV, v_off_fin = 2948.289 uV 09:35:31 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -160.853 uVs 09:35:31 Integrated Voltage Measurement... 09:35:40 Voltage Offset Correction: v_off_ini = 2955.885 uV, v_off_fin = 2951.018 uV 09:35:40 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 171.686 uVs 09:35:40 Integrated Voltage Measurement... 09:35:49 Voltage Offset Correction: v_off_ini = 2938.339 uV, v_off_fin = 2938.645 uV 09:35:49 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -212.931 uVs 09:35:49 Integrated Voltage Measurement... 09:35:58 Voltage Offset Correction: v_off_ini = 2942.878 uV, v_off_fin = 2943.839 uV 09:35:58 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 161.312 uVs 09:36:00 I2X Average and RMS of Measurements VT = -180.266 +- 9.898 uVs I2X = 16.584 +- 0.911 uTm^2 09:36:05 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:36:08 Integrated Voltage Measurement... 09:36:17 Voltage Offset Correction: v_off_ini = 2940.811 uV, v_off_fin = 2937.222 uV 09:36:17 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -320.291 uVs 09:36:17 Integrated Voltage Measurement... 09:36:26 Voltage Offset Correction: v_off_ini = 2939.964 uV, v_off_fin = 2946.548 uV 09:36:26 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 294.501 uVs 09:36:26 Integrated Voltage Measurement... 09:36:35 Voltage Offset Correction: v_off_ini = 2937.054 uV, v_off_fin = 2941.399 uV 09:36:35 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -310.518 uVs 09:36:35 Integrated Voltage Measurement... 09:36:44 Voltage Offset Correction: v_off_ini = 2942.826 uV, v_off_fin = 2922.631 uV 09:36:44 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 293.478 uVs 09:36:44 Integrated Voltage Measurement... 09:36:53 Voltage Offset Correction: v_off_ini = 2948.635 uV, v_off_fin = 2942.373 uV 09:36:53 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -319.019 uVs 09:36:53 Integrated Voltage Measurement... 09:37:02 Voltage Offset Correction: v_off_ini = 2960.330 uV, v_off_fin = 2975.559 uV 09:37:02 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 298.580 uVs 09:37:05 I2Y Average and RMS of Measurements VT = -306.064 +- 2.932 uVs I2Y = -28.158 +- 0.270 uTm^2 09:37:19 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:37:23 Integrated Voltage Measurement... 09:37:32 Voltage Offset Correction: v_off_ini = 2968.593 uV, v_off_fin = 2964.198 uV 09:37:32 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.4998 mm dXY = -0.4999 mm VT = -186.847 uVs 09:37:32 Integrated Voltage Measurement... 09:37:41 Voltage Offset Correction: v_off_ini = 2975.427 uV, v_off_fin = 2991.230 uV 09:37:41 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 189.470 uVs 09:37:41 Integrated Voltage Measurement... 09:37:50 Voltage Offset Correction: v_off_ini = 2968.264 uV, v_off_fin = 2972.606 uV 09:37:50 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -160.966 uVs 09:37:50 Integrated Voltage Measurement... 09:37:59 Voltage Offset Correction: v_off_ini = 2973.128 uV, v_off_fin = 2978.122 uV 09:37:59 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 161.870 uVs 09:37:59 Integrated Voltage Measurement... 09:38:08 Voltage Offset Correction: v_off_ini = 2978.806 uV, v_off_fin = 2984.138 uV 09:38:08 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -147.967 uVs 09:38:08 Integrated Voltage Measurement... 09:38:17 Voltage Offset Correction: v_off_ini = 2988.343 uV, v_off_fin = 2983.859 uV 09:38:17 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 156.532 uVs 09:38:22 I1X Average and RMS of Measurements VT = -167.275 +- 15.234 uVs I1X = 33.457 +- 3.049 uTm 09:38:24 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:38:28 Integrated Voltage Measurement... 09:38:37 Voltage Offset Correction: v_off_ini = 2960.705 uV, v_off_fin = 2975.337 uV 09:38:37 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -242.589 uVs 09:38:37 Integrated Voltage Measurement... 09:38:46 Voltage Offset Correction: v_off_ini = 2984.630 uV, v_off_fin = 2993.057 uV 09:38:46 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 256.378 uVs 09:38:46 Integrated Voltage Measurement... 09:38:55 Voltage Offset Correction: v_off_ini = 2963.390 uV, v_off_fin = 2968.251 uV 09:38:55 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -251.678 uVs 09:38:55 Integrated Voltage Measurement... 09:39:04 Voltage Offset Correction: v_off_ini = 2975.925 uV, v_off_fin = 3004.250 uV 09:39:04 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 259.421 uVs 09:39:05 Integrated Voltage Measurement... 09:39:14 Voltage Offset Correction: v_off_ini = 2971.157 uV, v_off_fin = 2975.429 uV 09:39:14 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -243.979 uVs 09:39:14 Integrated Voltage Measurement... 09:39:23 Voltage Offset Correction: v_off_ini = 3004.247 uV, v_off_fin = 3017.708 uV 09:39:23 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 244.603 uVs 09:39:27 I1Y Average and RMS of Measurements VT = -249.775 +- 4.601 uVs I1Y = -49.950 +- 0.920 uTm 09:39:32 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:39:35 Integrated Voltage Measurement... 09:39:44 Voltage Offset Correction: v_off_ini = 2983.960 uV, v_off_fin = 2991.814 uV 09:39:44 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -146.377 uVs 09:39:44 Integrated Voltage Measurement... 09:39:53 Voltage Offset Correction: v_off_ini = 3006.094 uV, v_off_fin = 2997.483 uV 09:39:53 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 165.821 uVs 09:39:53 Integrated Voltage Measurement... 09:40:02 Voltage Offset Correction: v_off_ini = 2992.272 uV, v_off_fin = 2974.140 uV 09:40:02 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -196.713 uVs 09:40:02 Integrated Voltage Measurement... 09:40:11 Voltage Offset Correction: v_off_ini = 2989.952 uV, v_off_fin = 2986.724 uV 09:40:11 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 183.195 uVs 09:40:11 Integrated Voltage Measurement... 09:40:20 Voltage Offset Correction: v_off_ini = 2991.779 uV, v_off_fin = 2978.242 uV 09:40:20 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -185.953 uVs 09:40:20 Integrated Voltage Measurement... 09:40:30 Voltage Offset Correction: v_off_ini = 3003.681 uV, v_off_fin = 3004.597 uV 09:40:30 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 184.068 uVs 09:40:32 I2X Average and RMS of Measurements VT = -177.021 +- 14.931 uVs I2X = 16.286 +- 1.374 uTm^2 09:40:37 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:40:40 Integrated Voltage Measurement... 09:40:49 Voltage Offset Correction: v_off_ini = 2990.848 uV, v_off_fin = 2987.054 uV 09:40:49 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -264.346 uVs 09:40:49 Integrated Voltage Measurement... 09:40:58 Voltage Offset Correction: v_off_ini = 2992.019 uV, v_off_fin = 2977.125 uV 09:40:58 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 288.325 uVs 09:40:58 Integrated Voltage Measurement... 09:41:07 Voltage Offset Correction: v_off_ini = 2999.062 uV, v_off_fin = 2989.071 uV 09:41:07 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -289.789 uVs 09:41:07 Integrated Voltage Measurement... 09:41:16 Voltage Offset Correction: v_off_ini = 2995.011 uV, v_off_fin = 3003.727 uV 09:41:16 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 293.249 uVs 09:41:16 Integrated Voltage Measurement... 09:41:25 Voltage Offset Correction: v_off_ini = 3017.861 uV, v_off_fin = 3001.410 uV 09:41:25 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -250.677 uVs 09:41:26 Integrated Voltage Measurement... 09:41:35 Voltage Offset Correction: v_off_ini = 3022.409 uV, v_off_fin = 3000.201 uV 09:41:35 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 284.237 uVs 09:41:37 I2Y Average and RMS of Measurements VT = -278.437 +- 9.935 uVs I2Y = -25.616 +- 0.914 uTm^2 09:41:51 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:41:56 Integrated Voltage Measurement... 09:42:05 Voltage Offset Correction: v_off_ini = 3014.717 uV, v_off_fin = 3003.698 uV 09:42:05 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.4998 mm dXY = -0.4999 mm VT = -184.354 uVs 09:42:05 Integrated Voltage Measurement... 09:42:14 Voltage Offset Correction: v_off_ini = 2996.859 uV, v_off_fin = 2999.422 uV 09:42:14 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 192.602 uVs 09:42:14 Integrated Voltage Measurement... 09:42:23 Voltage Offset Correction: v_off_ini = 3001.013 uV, v_off_fin = 3010.210 uV 09:42:23 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -193.275 uVs 09:42:23 Integrated Voltage Measurement... 09:42:32 Voltage Offset Correction: v_off_ini = 2986.476 uV, v_off_fin = 3007.771 uV 09:42:32 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 149.259 uVs 09:42:32 Integrated Voltage Measurement... 09:42:41 Voltage Offset Correction: v_off_ini = 3008.043 uV, v_off_fin = 2989.043 uV 09:42:41 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -177.072 uVs 09:42:41 Integrated Voltage Measurement... 09:42:50 Voltage Offset Correction: v_off_ini = 3014.261 uV, v_off_fin = 3008.606 uV 09:42:50 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 161.737 uVs 09:42:55 I1X Average and RMS of Measurements VT = -176.383 +- 8.586 uVs I1X = 35.278 +- 1.720 uTm 09:42:57 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:43:02 Integrated Voltage Measurement... 09:43:11 Voltage Offset Correction: v_off_ini = 3015.870 uV, v_off_fin = 2989.257 uV 09:43:11 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -246.972 uVs 09:43:11 Integrated Voltage Measurement... 09:43:20 Voltage Offset Correction: v_off_ini = 2999.948 uV, v_off_fin = 2996.469 uV 09:43:20 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 261.167 uVs 09:43:20 Integrated Voltage Measurement... 09:43:29 Voltage Offset Correction: v_off_ini = 2975.975 uV, v_off_fin = 2969.804 uV 09:43:29 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -224.281 uVs 09:43:29 Integrated Voltage Measurement... 09:43:38 Voltage Offset Correction: v_off_ini = 2994.087 uV, v_off_fin = 3019.171 uV 09:43:38 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 258.729 uVs 09:43:38 Integrated Voltage Measurement... 09:43:47 Voltage Offset Correction: v_off_ini = 3011.451 uV, v_off_fin = 2973.324 uV 09:43:47 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -255.883 uVs 09:43:47 Integrated Voltage Measurement... 09:43:56 Voltage Offset Correction: v_off_ini = 3000.386 uV, v_off_fin = 2995.308 uV 09:43:56 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 229.029 uVs 09:44:00 I1Y Average and RMS of Measurements VT = -246.010 +- 5.712 uVs I1Y = -49.199 +- 1.141 uTm 09:44:05 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:44:08 Integrated Voltage Measurement... 09:44:17 Voltage Offset Correction: v_off_ini = 2966.232 uV, v_off_fin = 2978.844 uV 09:44:17 Integrated Voltage Measurement dXYus = -1.0001 mm, dXYds = 0.0000 mm dXY = -1.0001 mm VT = -151.845 uVs 09:44:17 Integrated Voltage Measurement... 09:44:26 Voltage Offset Correction: v_off_ini = 2984.895 uV, v_off_fin = 3001.162 uV 09:44:26 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 154.277 uVs 09:44:26 Integrated Voltage Measurement... 09:44:35 Voltage Offset Correction: v_off_ini = 2985.600 uV, v_off_fin = 2990.765 uV 09:44:35 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -192.854 uVs 09:44:35 Integrated Voltage Measurement... 09:44:44 Voltage Offset Correction: v_off_ini = 2984.975 uV, v_off_fin = 3001.877 uV 09:44:44 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 178.544 uVs 09:44:44 Integrated Voltage Measurement... 09:44:53 Voltage Offset Correction: v_off_ini = 2968.116 uV, v_off_fin = 2958.945 uV 09:44:53 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -170.508 uVs 09:44:53 Integrated Voltage Measurement... 09:45:02 Voltage Offset Correction: v_off_ini = 2986.837 uV, v_off_fin = 2997.403 uV 09:45:02 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 161.553 uVs 09:45:04 I2X Average and RMS of Measurements VT = -168.263 +- 13.418 uVs I2X = 15.480 +- 1.235 uTm^2 09:45:10 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:45:12 Integrated Voltage Measurement... 09:45:21 Voltage Offset Correction: v_off_ini = 2942.197 uV, v_off_fin = 2932.590 uV 09:45:21 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -273.614 uVs 09:45:21 Integrated Voltage Measurement... 09:45:30 Voltage Offset Correction: v_off_ini = 2960.911 uV, v_off_fin = 2956.970 uV 09:45:30 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 274.636 uVs 09:45:30 Integrated Voltage Measurement... 09:45:39 Voltage Offset Correction: v_off_ini = 2928.388 uV, v_off_fin = 2937.004 uV 09:45:39 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -280.015 uVs 09:45:39 Integrated Voltage Measurement... 09:45:48 Voltage Offset Correction: v_off_ini = 2934.922 uV, v_off_fin = 2944.034 uV 09:45:48 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 272.738 uVs 09:45:48 Integrated Voltage Measurement... 09:45:57 Voltage Offset Correction: v_off_ini = 2931.342 uV, v_off_fin = 2935.873 uV 09:45:57 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -300.436 uVs 09:45:57 Integrated Voltage Measurement... 09:46:06 Voltage Offset Correction: v_off_ini = 2941.084 uV, v_off_fin = 2969.933 uV 09:46:06 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 279.251 uVs 09:46:09 I2Y Average and RMS of Measurements VT = -280.115 +- 6.940 uVs I2Y = -25.771 +- 0.639 uTm^2 09:46:23 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:46:27 Integrated Voltage Measurement... 09:46:36 Voltage Offset Correction: v_off_ini = 2937.017 uV, v_off_fin = 2926.417 uV 09:46:36 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -156.907 uVs 09:46:36 Integrated Voltage Measurement... 09:46:45 Voltage Offset Correction: v_off_ini = 2941.361 uV, v_off_fin = 2943.104 uV 09:46:45 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 194.619 uVs 09:46:45 Integrated Voltage Measurement... 09:46:54 Voltage Offset Correction: v_off_ini = 2937.038 uV, v_off_fin = 2944.322 uV 09:46:54 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -158.942 uVs 09:46:54 Integrated Voltage Measurement... 09:47:03 Voltage Offset Correction: v_off_ini = 2943.257 uV, v_off_fin = 2954.644 uV 09:47:03 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 150.117 uVs 09:47:03 Integrated Voltage Measurement... 09:47:12 Voltage Offset Correction: v_off_ini = 2933.940 uV, v_off_fin = 2959.690 uV 09:47:12 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -129.961 uVs 09:47:12 Integrated Voltage Measurement... 09:47:22 Voltage Offset Correction: v_off_ini = 2955.547 uV, v_off_fin = 2944.020 uV 09:47:22 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 173.657 uVs 09:47:26 I1X Average and RMS of Measurements VT = -160.701 +- 10.709 uVs I1X = 32.141 +- 2.141 uTm 09:47:28 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:47:32 Integrated Voltage Measurement... 09:47:41 Voltage Offset Correction: v_off_ini = 2948.976 uV, v_off_fin = 2935.326 uV 09:47:41 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -206.955 uVs 09:47:41 Integrated Voltage Measurement... 09:47:50 Voltage Offset Correction: v_off_ini = 2945.894 uV, v_off_fin = 2949.691 uV 09:47:50 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 236.393 uVs 09:47:51 Integrated Voltage Measurement... 09:48:00 Voltage Offset Correction: v_off_ini = 2925.253 uV, v_off_fin = 2916.809 uV 09:48:00 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -220.611 uVs 09:48:00 Integrated Voltage Measurement... 09:48:09 Voltage Offset Correction: v_off_ini = 2944.928 uV, v_off_fin = 2959.637 uV 09:48:09 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 243.179 uVs 09:48:09 Integrated Voltage Measurement... 09:48:18 Voltage Offset Correction: v_off_ini = 2953.988 uV, v_off_fin = 2926.851 uV 09:48:18 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -210.614 uVs 09:48:18 Integrated Voltage Measurement... 09:48:27 Voltage Offset Correction: v_off_ini = 2928.616 uV, v_off_fin = 2925.101 uV 09:48:27 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 239.322 uVs 09:48:31 I1Y Average and RMS of Measurements VT = -226.179 +- 4.260 uVs I1Y = -45.232 +- 0.851 uTm 09:48:37 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:48:39 Integrated Voltage Measurement... 09:48:48 Voltage Offset Correction: v_off_ini = 2926.477 uV, v_off_fin = 2902.823 uV 09:48:48 Integrated Voltage Measurement dXYus = -1.0001 mm, dXYds = 0.0000 mm dXY = -1.0001 mm VT = -179.363 uVs 09:48:48 Integrated Voltage Measurement... 09:48:57 Voltage Offset Correction: v_off_ini = 2919.505 uV, v_off_fin = 2946.354 uV 09:48:57 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 158.288 uVs 09:48:57 Integrated Voltage Measurement... 09:49:06 Voltage Offset Correction: v_off_ini = 2923.906 uV, v_off_fin = 2902.503 uV 09:49:06 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -194.064 uVs 09:49:06 Integrated Voltage Measurement... 09:49:15 Voltage Offset Correction: v_off_ini = 2904.002 uV, v_off_fin = 2921.473 uV 09:49:15 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 161.476 uVs 09:49:16 Integrated Voltage Measurement... 09:49:25 Voltage Offset Correction: v_off_ini = 2922.579 uV, v_off_fin = 2917.856 uV 09:49:25 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -192.639 uVs 09:49:25 Integrated Voltage Measurement... 09:49:34 Voltage Offset Correction: v_off_ini = 2928.210 uV, v_off_fin = 2951.381 uV 09:49:34 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 149.551 uVs 09:49:36 I2X Average and RMS of Measurements VT = -172.564 +- 3.796 uVs I2X = 15.876 +- 0.350 uTm^2 09:49:41 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:49:44 Integrated Voltage Measurement... 09:49:53 Voltage Offset Correction: v_off_ini = 2918.058 uV, v_off_fin = 2903.327 uV 09:49:53 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -233.882 uVs 09:49:53 Integrated Voltage Measurement... 09:50:02 Voltage Offset Correction: v_off_ini = 2888.805 uV, v_off_fin = 2913.514 uV 09:50:02 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 244.162 uVs 09:50:02 Integrated Voltage Measurement... 09:50:11 Voltage Offset Correction: v_off_ini = 2932.303 uV, v_off_fin = 2922.173 uV 09:50:11 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -245.496 uVs 09:50:11 Integrated Voltage Measurement... 09:50:20 Voltage Offset Correction: v_off_ini = 2906.381 uV, v_off_fin = 2912.004 uV 09:50:20 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 261.639 uVs 09:50:20 Integrated Voltage Measurement... 09:50:29 Voltage Offset Correction: v_off_ini = 2897.622 uV, v_off_fin = 2907.073 uV 09:50:29 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -233.199 uVs 09:50:29 Integrated Voltage Measurement... 09:50:38 Voltage Offset Correction: v_off_ini = 2913.320 uV, v_off_fin = 2921.654 uV 09:50:38 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 256.498 uVs 09:50:41 I2Y Average and RMS of Measurements VT = -245.813 +- 5.977 uVs I2Y = -22.615 +- 0.550 uTm^2 09:50:43 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm