SLAC Magnetic Measurements Date: 08-13-2026 Time: 08:20:46 Project: LCLS-II-HE Magnet Type: Phase Shifter Magnet Name: PS_DELTA-II Dataset Number: 1 Run Purpose: Tuning Run Type: Run Number: 12 Measurement Device: Stretched Wire Operator: YL Comment: 08:20:53 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 08:20:56 Motion Parameters, X Controller, Axis 1 acc = 0.001000 m/s/s dcl = 0.001000 m/s/s vel = 0.001000 m/s pos = 0.000000 m kp = 188.600000 ki = 820.000000 kd = 491.400000 db = 0.000000 08:20:57 Motion Parameters, X Controller, Axis 2 acc = 0.001000 m/s/s dcl = 0.001000 m/s/s vel = 0.001000 m/s pos = 0.000000 m kp = 188.600000 ki = 820.000000 kd = 491.400000 db = 0.000000 08:20:59 Motion Parameters, Y Controller, Axis 1 acc = 0.001000 m/s/s dcl = 0.001000 m/s/s vel = 0.001000 m/s pos = 0.000000 m kp = 188.600000 ki = 820.000000 kd = 491.400000 db = 0.000000 08:21:00 Motion Parameters, Y Controller, Axis 2 acc = 0.001000 m/s/s dcl = 0.001000 m/s/s vel = 0.001000 m/s pos = 0.000000 m kp = 188.600000 ki = 820.000000 kd = 491.400000 db = 0.000000 08:21:11 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 08:21:15 Integrated Voltage Measurement... 08:21:24 Voltage Offset Correction: v_off_ini = 2951.702 uV, v_off_fin = 2949.766 uV 08:21:24 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.4998 mm dXY = -0.4999 mm VT = -106.887 uVs 08:21:24 Integrated Voltage Measurement... 08:21:33 Voltage Offset Correction: v_off_ini = 2943.146 uV, v_off_fin = 2927.848 uV 08:21:33 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 124.060 uVs 08:21:34 Integrated Voltage Measurement... 08:21:42 Voltage Offset Correction: v_off_ini = 2920.233 uV, v_off_fin = 2926.589 uV 08:21:42 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -95.996 uVs 08:21:43 Integrated Voltage Measurement... 08:21:51 Voltage Offset Correction: v_off_ini = 2934.683 uV, v_off_fin = 2924.386 uV 08:21:51 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 78.532 uVs 08:21:52 Integrated Voltage Measurement... 08:22:00 Voltage Offset Correction: v_off_ini = 2936.171 uV, v_off_fin = 2933.765 uV 08:22:00 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -117.926 uVs 08:22:01 Integrated Voltage Measurement... 08:22:09 Voltage Offset Correction: v_off_ini = 2927.364 uV, v_off_fin = 2933.385 uV 08:22:09 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 105.887 uVs 08:22:14 I1X Average and RMS of Measurements VT = -104.881 +- 12.542 uVs I1X = 20.977 +- 2.509 uTm 08:22:16 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 08:22:20 Integrated Voltage Measurement...