SLAC Magnetic Measurements Date: 07-17-2025 Time: 09:33:33 Project: LCLS-II-HE Magnet Type: Phase Shifter Magnet Name: HE_SXPS_16353 Dataset Number: 1 Run Purpose: Tuning Run Type: Stretched Wire Run Number: 2 Measurement Device: Operator: YL Comment: 09:34:31 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:34:34 Motion Parameters, X Controller, Axis 1 acc = 0.001000 m/s/s dcl = 0.001000 m/s/s vel = 0.001000 m/s pos = 0.000000 m kp = 188.600000 ki = 820.000000 kd = 491.400000 db = 0.000000 09:34:35 Motion Parameters, X Controller, Axis 2 acc = 0.001000 m/s/s dcl = 0.001000 m/s/s vel = 0.001000 m/s pos = 0.000000 m kp = 188.600000 ki = 820.000000 kd = 491.400000 db = 0.000000 09:34:37 Motion Parameters, Y Controller, Axis 1 acc = 0.001000 m/s/s dcl = 0.001000 m/s/s vel = 0.001000 m/s pos = 0.000000 m kp = 188.600000 ki = 820.000000 kd = 491.400000 db = 0.000000 09:34:39 Motion Parameters, Y Controller, Axis 2 acc = 0.001000 m/s/s dcl = 0.001000 m/s/s vel = 0.001000 m/s pos = 0.000000 m kp = 188.600000 ki = 820.000000 kd = 491.400000 db = 0.000000 09:43:40 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:43:44 Integrated Voltage Measurement... 09:43:53 Voltage Offset Correction: v_off_ini = 284.593 uV, v_off_fin = 262.963 uV 09:43:53 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.4997 mm dXY = -0.4998 mm VT = -101.317 uVs 09:43:53 Integrated Voltage Measurement... 09:44:02 Voltage Offset Correction: v_off_ini = 271.926 uV, v_off_fin = 270.750 uV 09:44:02 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = 103.226 uVs 09:44:02 Integrated Voltage Measurement... 09:44:11 Voltage Offset Correction: v_off_ini = 268.703 uV, v_off_fin = 259.874 uV 09:44:11 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -108.521 uVs 09:44:11 Integrated Voltage Measurement... 09:44:20 Voltage Offset Correction: v_off_ini = 262.008 uV, v_off_fin = 271.551 uV 09:44:20 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = 97.127 uVs 09:44:21 Integrated Voltage Measurement... 09:44:29 Voltage Offset Correction: v_off_ini = 271.883 uV, v_off_fin = 267.581 uV 09:44:30 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -128.096 uVs 09:44:30 Integrated Voltage Measurement... 09:44:39 Voltage Offset Correction: v_off_ini = 267.791 uV, v_off_fin = 276.687 uV 09:44:39 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 112.227 uVs 09:44:43 I1X Average and RMS of Measurements VT = -108.419 +- 8.306 uVs I1X = 21.686 +- 1.660 uTm 09:44:45 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:44:50 Integrated Voltage Measurement... 09:44:59 Voltage Offset Correction: v_off_ini = 285.453 uV, v_off_fin = 263.120 uV 09:44:59 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -193.145 uVs 09:44:59 Integrated Voltage Measurement... 09:45:08 Voltage Offset Correction: v_off_ini = 266.653 uV, v_off_fin = 281.922 uV 09:45:08 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 200.853 uVs 09:45:08 Integrated Voltage Measurement... 09:45:17 Voltage Offset Correction: v_off_ini = 267.809 uV, v_off_fin = 256.876 uV 09:45:17 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -188.969 uVs 09:45:17 Integrated Voltage Measurement... 09:45:26 Voltage Offset Correction: v_off_ini = 256.246 uV, v_off_fin = 281.531 uV 09:45:26 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 197.219 uVs 09:45:26 Integrated Voltage Measurement... 09:45:35 Voltage Offset Correction: v_off_ini = 269.237 uV, v_off_fin = 260.575 uV 09:45:35 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -171.898 uVs 09:45:35 Integrated Voltage Measurement... 09:45:44 Voltage Offset Correction: v_off_ini = 269.189 uV, v_off_fin = 281.100 uV 09:45:44 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 183.607 uVs 09:45:48 I1Y Average and RMS of Measurements VT = -189.282 +- 8.307 uVs I1Y = -37.856 +- 1.661 uTm 09:45:53 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:45:55 Integrated Voltage Measurement... 09:46:04 Voltage Offset Correction: v_off_ini = 263.361 uV, v_off_fin = 265.931 uV 09:46:04 Integrated Voltage Measurement dXYus = -1.0002 mm, dXYds = 0.0000 mm dXY = -1.0002 mm VT = -91.427 uVs 09:46:05 Integrated Voltage Measurement... 09:46:14 Voltage Offset Correction: v_off_ini = 272.682 uV, v_off_fin = 279.740 uV 09:46:14 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 77.227 uVs 09:46:14 Integrated Voltage Measurement... 09:46:23 Voltage Offset Correction: v_off_ini = 268.621 uV, v_off_fin = 270.664 uV 09:46:23 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -83.909 uVs 09:46:23 Integrated Voltage Measurement... 09:46:32 Voltage Offset Correction: v_off_ini = 284.154 uV, v_off_fin = 290.147 uV 09:46:32 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 64.461 uVs 09:46:32 Integrated Voltage Measurement... 09:46:41 Voltage Offset Correction: v_off_ini = 287.055 uV, v_off_fin = 273.638 uV 09:46:41 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -76.090 uVs 09:46:41 Integrated Voltage Measurement... 09:46:50 Voltage Offset Correction: v_off_ini = 282.993 uV, v_off_fin = 301.630 uV 09:46:50 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 68.115 uVs 09:46:52 I2X Average and RMS of Measurements VT = -76.871 +- 5.340 uVs I2X = 7.072 +- 0.491 uTm^2 09:46:57 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:47:00 Integrated Voltage Measurement... 09:47:09 Voltage Offset Correction: v_off_ini = 278.783 uV, v_off_fin = 274.307 uV 09:47:09 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -523.091 uVs 09:47:09 Integrated Voltage Measurement... 09:47:18 Voltage Offset Correction: v_off_ini = 285.836 uV, v_off_fin = 292.971 uV 09:47:18 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 510.037 uVs 09:47:18 Integrated Voltage Measurement... 09:47:27 Voltage Offset Correction: v_off_ini = 290.785 uV, v_off_fin = 294.013 uV 09:47:27 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -541.700 uVs 09:47:27 Integrated Voltage Measurement... 09:47:36 Voltage Offset Correction: v_off_ini = 297.441 uV, v_off_fin = 288.414 uV 09:47:36 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 529.586 uVs 09:47:36 Integrated Voltage Measurement... 09:47:45 Voltage Offset Correction: v_off_ini = 295.301 uV, v_off_fin = 290.401 uV 09:47:45 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -521.528 uVs 09:47:45 Integrated Voltage Measurement... 09:47:54 Voltage Offset Correction: v_off_ini = 285.502 uV, v_off_fin = 295.215 uV 09:47:54 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 531.909 uVs 09:47:56 I2Y Average and RMS of Measurements VT = -526.308 +- 7.794 uVs I2Y = -48.420 +- 0.717 uTm^2 09:48:08 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:48:13 Integrated Voltage Measurement... 09:48:21 Voltage Offset Correction: v_off_ini = 304.528 uV, v_off_fin = 300.232 uV 09:48:22 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.4997 mm dXY = -0.4998 mm VT = -145.216 uVs 09:48:22 Integrated Voltage Measurement... 09:48:31 Voltage Offset Correction: v_off_ini = 285.241 uV, v_off_fin = 284.113 uV 09:48:31 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 125.876 uVs 09:48:31 Integrated Voltage Measurement... 09:48:40 Voltage Offset Correction: v_off_ini = 285.886 uV, v_off_fin = 278.827 uV 09:48:40 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -137.697 uVs 09:48:40 Integrated Voltage Measurement... 09:48:49 Voltage Offset Correction: v_off_ini = 289.633 uV, v_off_fin = 276.823 uV 09:48:49 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 122.279 uVs 09:48:49 Integrated Voltage Measurement... 09:48:58 Voltage Offset Correction: v_off_ini = 282.480 uV, v_off_fin = 289.890 uV 09:48:58 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -122.959 uVs 09:48:58 Integrated Voltage Measurement... 09:49:07 Voltage Offset Correction: v_off_ini = 292.392 uV, v_off_fin = 287.676 uV 09:49:07 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 109.553 uVs 09:49:11 I1X Average and RMS of Measurements VT = -127.264 +- 8.107 uVs I1X = 25.455 +- 1.623 uTm 09:49:13 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:49:18 Integrated Voltage Measurement... 09:49:27 Voltage Offset Correction: v_off_ini = 280.083 uV, v_off_fin = 272.201 uV 09:49:27 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -185.316 uVs 09:49:27 Integrated Voltage Measurement... 09:49:36 Voltage Offset Correction: v_off_ini = 295.804 uV, v_off_fin = 306.923 uV 09:49:36 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 187.045 uVs 09:49:36 Integrated Voltage Measurement... 09:49:45 Voltage Offset Correction: v_off_ini = 289.509 uV, v_off_fin = 277.428 uV 09:49:45 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -198.206 uVs 09:49:45 Integrated Voltage Measurement... 09:49:54 Voltage Offset Correction: v_off_ini = 279.648 uV, v_off_fin = 296.633 uV 09:49:54 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 208.783 uVs 09:49:54 Integrated Voltage Measurement... 09:50:03 Voltage Offset Correction: v_off_ini = 280.748 uV, v_off_fin = 266.994 uV 09:50:03 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -190.086 uVs 09:50:03 Integrated Voltage Measurement... 09:50:12 Voltage Offset Correction: v_off_ini = 288.692 uV, v_off_fin = 302.732 uV 09:50:12 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 196.653 uVs 09:50:16 I1Y Average and RMS of Measurements VT = -194.348 +- 7.102 uVs I1Y = -38.870 +- 1.420 uTm 09:50:21 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:50:24 Integrated Voltage Measurement... 09:50:33 Voltage Offset Correction: v_off_ini = 297.661 uV, v_off_fin = 285.134 uV 09:50:33 Integrated Voltage Measurement dXYus = -1.0002 mm, dXYds = 0.0000 mm dXY = -1.0002 mm VT = -89.595 uVs 09:50:33 Integrated Voltage Measurement... 09:50:42 Voltage Offset Correction: v_off_ini = 294.717 uV, v_off_fin = 318.258 uV 09:50:42 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 93.801 uVs 09:50:42 Integrated Voltage Measurement... 09:50:51 Voltage Offset Correction: v_off_ini = 299.065 uV, v_off_fin = 296.681 uV 09:50:51 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -70.558 uVs 09:50:51 Integrated Voltage Measurement... 09:51:00 Voltage Offset Correction: v_off_ini = 308.649 uV, v_off_fin = 316.612 uV 09:51:00 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 99.491 uVs 09:51:00 Integrated Voltage Measurement... 09:51:09 Voltage Offset Correction: v_off_ini = 301.212 uV, v_off_fin = 294.294 uV 09:51:09 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -98.355 uVs 09:51:09 Integrated Voltage Measurement... 09:51:18 Voltage Offset Correction: v_off_ini = 308.010 uV, v_off_fin = 310.601 uV 09:51:18 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 93.985 uVs 09:51:20 I2X Average and RMS of Measurements VT = -90.964 +- 4.580 uVs I2X = 8.368 +- 0.421 uTm^2 09:51:25 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:51:28 Integrated Voltage Measurement... 09:51:37 Voltage Offset Correction: v_off_ini = 310.901 uV, v_off_fin = 305.171 uV 09:51:37 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -501.709 uVs 09:51:37 Integrated Voltage Measurement... 09:51:46 Voltage Offset Correction: v_off_ini = 318.709 uV, v_off_fin = 320.069 uV 09:51:46 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 523.601 uVs 09:51:46 Integrated Voltage Measurement... 09:51:55 Voltage Offset Correction: v_off_ini = 322.424 uV, v_off_fin = 322.335 uV 09:51:55 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -521.947 uVs 09:51:55 Integrated Voltage Measurement... 09:52:04 Voltage Offset Correction: v_off_ini = 317.066 uV, v_off_fin = 314.942 uV 09:52:04 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 527.423 uVs 09:52:04 Integrated Voltage Measurement... 09:52:13 Voltage Offset Correction: v_off_ini = 320.859 uV, v_off_fin = 312.225 uV 09:52:13 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -528.300 uVs 09:52:13 Integrated Voltage Measurement... 09:52:22 Voltage Offset Correction: v_off_ini = 307.658 uV, v_off_fin = 317.961 uV 09:52:22 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 536.677 uVs 09:52:25 I2Y Average and RMS of Measurements VT = -523.276 +- 8.158 uVs I2Y = -48.141 +- 0.751 uTm^2 09:52:36 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:52:41 Integrated Voltage Measurement... 09:52:50 Voltage Offset Correction: v_off_ini = 320.847 uV, v_off_fin = 315.405 uV 09:52:50 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -149.967 uVs 09:52:50 Integrated Voltage Measurement... 09:52:59 Voltage Offset Correction: v_off_ini = 316.707 uV, v_off_fin = 314.542 uV 09:52:59 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 137.402 uVs 09:52:59 Integrated Voltage Measurement... 09:53:08 Voltage Offset Correction: v_off_ini = 329.020 uV, v_off_fin = 319.787 uV 09:53:08 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -126.032 uVs 09:53:08 Integrated Voltage Measurement... 09:53:17 Voltage Offset Correction: v_off_ini = 321.948 uV, v_off_fin = 322.767 uV 09:53:17 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 147.737 uVs 09:53:17 Integrated Voltage Measurement... 09:53:26 Voltage Offset Correction: v_off_ini = 321.330 uV, v_off_fin = 327.635 uV 09:53:26 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -130.922 uVs 09:53:26 Integrated Voltage Measurement... 09:53:35 Voltage Offset Correction: v_off_ini = 324.022 uV, v_off_fin = 327.308 uV 09:53:35 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 122.755 uVs 09:53:39 I1X Average and RMS of Measurements VT = -135.803 +- 6.920 uVs I1X = 27.161 +- 1.384 uTm 09:53:42 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:53:46 Integrated Voltage Measurement... 09:53:55 Voltage Offset Correction: v_off_ini = 325.221 uV, v_off_fin = 326.852 uV 09:53:55 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -202.129 uVs 09:53:55 Integrated Voltage Measurement... 09:54:04 Voltage Offset Correction: v_off_ini = 329.871 uV, v_off_fin = 345.397 uV 09:54:04 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 185.323 uVs 09:54:04 Integrated Voltage Measurement... 09:54:13 Voltage Offset Correction: v_off_ini = 328.689 uV, v_off_fin = 307.969 uV 09:54:13 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -203.757 uVs 09:54:13 Integrated Voltage Measurement... 09:54:22 Voltage Offset Correction: v_off_ini = 322.561 uV, v_off_fin = 327.358 uV 09:54:22 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 188.929 uVs 09:54:22 Integrated Voltage Measurement... 09:54:31 Voltage Offset Correction: v_off_ini = 316.728 uV, v_off_fin = 313.288 uV 09:54:31 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -188.504 uVs 09:54:31 Integrated Voltage Measurement... 09:54:40 Voltage Offset Correction: v_off_ini = 306.745 uV, v_off_fin = 316.299 uV 09:54:40 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 197.396 uVs 09:54:44 I1Y Average and RMS of Measurements VT = -194.340 +- 1.452 uVs I1Y = -38.868 +- 0.290 uTm 09:54:49 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:54:52 Integrated Voltage Measurement... 09:55:01 Voltage Offset Correction: v_off_ini = 312.677 uV, v_off_fin = 304.261 uV 09:55:01 Integrated Voltage Measurement dXYus = -1.0002 mm, dXYds = 0.0000 mm dXY = -1.0002 mm VT = -106.544 uVs 09:55:01 Integrated Voltage Measurement... 09:55:10 Voltage Offset Correction: v_off_ini = 311.975 uV, v_off_fin = 314.753 uV 09:55:10 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 105.443 uVs 09:55:10 Integrated Voltage Measurement... 09:55:19 Voltage Offset Correction: v_off_ini = 310.752 uV, v_off_fin = 298.468 uV 09:55:19 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -109.337 uVs 09:55:19 Integrated Voltage Measurement... 09:55:28 Voltage Offset Correction: v_off_ini = 304.546 uV, v_off_fin = 322.486 uV 09:55:28 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 93.058 uVs 09:55:28 Integrated Voltage Measurement... 09:55:37 Voltage Offset Correction: v_off_ini = 316.105 uV, v_off_fin = 296.300 uV 09:55:37 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -99.674 uVs 09:55:37 Integrated Voltage Measurement... 09:55:46 Voltage Offset Correction: v_off_ini = 300.809 uV, v_off_fin = 314.655 uV 09:55:46 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 112.387 uVs 09:55:48 I2X Average and RMS of Measurements VT = -104.407 +- 2.270 uVs I2X = 9.605 +- 0.209 uTm^2 09:55:54 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:55:56 Integrated Voltage Measurement... 09:56:05 Voltage Offset Correction: v_off_ini = 306.151 uV, v_off_fin = 313.528 uV 09:56:05 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -514.723 uVs 09:56:05 Integrated Voltage Measurement... 09:56:14 Voltage Offset Correction: v_off_ini = 312.927 uV, v_off_fin = 326.163 uV 09:56:14 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 518.118 uVs 09:56:14 Integrated Voltage Measurement... 09:56:23 Voltage Offset Correction: v_off_ini = 319.351 uV, v_off_fin = 313.916 uV 09:56:23 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -533.492 uVs 09:56:23 Integrated Voltage Measurement... 09:56:32 Voltage Offset Correction: v_off_ini = 311.675 uV, v_off_fin = 328.006 uV 09:56:32 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 506.943 uVs 09:56:32 Integrated Voltage Measurement... 09:56:41 Voltage Offset Correction: v_off_ini = 318.259 uV, v_off_fin = 326.268 uV 09:56:41 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -525.037 uVs 09:56:41 Integrated Voltage Measurement... 09:56:50 Voltage Offset Correction: v_off_ini = 333.009 uV, v_off_fin = 336.447 uV 09:56:50 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 496.786 uVs 09:56:53 I2Y Average and RMS of Measurements VT = -515.850 +- 3.821 uVs I2Y = -47.458 +- 0.352 uTm^2 09:57:05 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:57:09 Integrated Voltage Measurement... 09:57:18 Voltage Offset Correction: v_off_ini = 334.844 uV, v_off_fin = 324.182 uV 09:57:18 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -149.915 uVs 09:57:18 Integrated Voltage Measurement... 09:57:27 Voltage Offset Correction: v_off_ini = 324.775 uV, v_off_fin = 338.934 uV 09:57:27 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 154.966 uVs 09:57:27 Integrated Voltage Measurement... 09:57:36 Voltage Offset Correction: v_off_ini = 340.017 uV, v_off_fin = 329.171 uV 09:57:36 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -140.877 uVs 09:57:36 Integrated Voltage Measurement... 09:57:45 Voltage Offset Correction: v_off_ini = 341.381 uV, v_off_fin = 327.225 uV 09:57:45 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 150.267 uVs 09:57:45 Integrated Voltage Measurement... 09:57:54 Voltage Offset Correction: v_off_ini = 333.368 uV, v_off_fin = 325.025 uV 09:57:54 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.4999 mm dXY = -0.4999 mm VT = -151.223 uVs 09:57:54 Integrated Voltage Measurement... 09:58:03 Voltage Offset Correction: v_off_ini = 332.306 uV, v_off_fin = 328.842 uV 09:58:03 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 133.389 uVs 09:58:07 I1X Average and RMS of Measurements VT = -146.773 +- 4.224 uVs I1X = 29.357 +- 0.844 uTm 09:58:10 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:58:14 Integrated Voltage Measurement... 09:58:23 Voltage Offset Correction: v_off_ini = 321.649 uV, v_off_fin = 313.821 uV 09:58:23 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = -186.949 uVs 09:58:23 Integrated Voltage Measurement... 09:58:32 Voltage Offset Correction: v_off_ini = 319.411 uV, v_off_fin = 337.737 uV 09:58:32 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 198.457 uVs 09:58:32 Integrated Voltage Measurement... 09:58:41 Voltage Offset Correction: v_off_ini = 331.628 uV, v_off_fin = 321.173 uV 09:58:41 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -188.490 uVs 09:58:41 Integrated Voltage Measurement... 09:58:50 Voltage Offset Correction: v_off_ini = 332.356 uV, v_off_fin = 334.639 uV 09:58:50 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 198.263 uVs 09:58:50 Integrated Voltage Measurement... 09:58:59 Voltage Offset Correction: v_off_ini = 314.511 uV, v_off_fin = 320.953 uV 09:58:59 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -205.618 uVs 09:58:59 Integrated Voltage Measurement... 09:59:08 Voltage Offset Correction: v_off_ini = 323.416 uV, v_off_fin = 329.280 uV 09:59:08 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 202.557 uVs 09:59:13 I1Y Average and RMS of Measurements VT = -196.722 +- 5.215 uVs I1Y = -39.345 +- 1.043 uTm 09:59:18 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:59:20 Integrated Voltage Measurement... 09:59:29 Voltage Offset Correction: v_off_ini = 330.983 uV, v_off_fin = 332.713 uV 09:59:29 Integrated Voltage Measurement dXYus = -1.0002 mm, dXYds = 0.0000 mm dXY = -1.0002 mm VT = -120.928 uVs 09:59:29 Integrated Voltage Measurement... 09:59:38 Voltage Offset Correction: v_off_ini = 336.875 uV, v_off_fin = 341.844 uV 09:59:38 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 115.285 uVs 09:59:38 Integrated Voltage Measurement... 09:59:47 Voltage Offset Correction: v_off_ini = 322.504 uV, v_off_fin = 325.911 uV 09:59:47 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -114.520 uVs 09:59:47 Integrated Voltage Measurement... 09:59:56 Voltage Offset Correction: v_off_ini = 322.866 uV, v_off_fin = 349.576 uV 09:59:56 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 126.977 uVs 09:59:56 Integrated Voltage Measurement... 10:00:05 Voltage Offset Correction: v_off_ini = 343.245 uV, v_off_fin = 330.825 uV 10:00:05 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -113.170 uVs 10:00:05 Integrated Voltage Measurement... 10:00:14 Voltage Offset Correction: v_off_ini = 341.858 uV, v_off_fin = 342.314 uV 10:00:14 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 138.766 uVs 10:00:17 I2X Average and RMS of Measurements VT = -121.608 +- 3.267 uVs I2X = 11.188 +- 0.301 uTm^2 10:00:22 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:00:25 Integrated Voltage Measurement... 10:00:34 Voltage Offset Correction: v_off_ini = 341.590 uV, v_off_fin = 339.193 uV 10:00:34 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -522.801 uVs 10:00:34 Integrated Voltage Measurement... 10:00:43 Voltage Offset Correction: v_off_ini = 346.401 uV, v_off_fin = 349.190 uV 10:00:43 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 511.056 uVs 10:00:43 Integrated Voltage Measurement... 10:00:52 Voltage Offset Correction: v_off_ini = 345.640 uV, v_off_fin = 339.131 uV 10:00:52 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -514.710 uVs 10:00:52 Integrated Voltage Measurement... 10:01:01 Voltage Offset Correction: v_off_ini = 332.705 uV, v_off_fin = 341.603 uV 10:01:01 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 500.630 uVs 10:01:01 Integrated Voltage Measurement... 10:01:10 Voltage Offset Correction: v_off_ini = 340.432 uV, v_off_fin = 348.387 uV 10:01:10 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -513.577 uVs 10:01:10 Integrated Voltage Measurement... 10:01:19 Voltage Offset Correction: v_off_ini = 355.026 uV, v_off_fin = 340.325 uV 10:01:19 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 491.472 uVs 10:01:21 I2Y Average and RMS of Measurements VT = -509.041 +- 5.960 uVs I2Y = -46.832 +- 0.548 uTm^2 10:01:33 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:01:37 Integrated Voltage Measurement... 10:01:46 Voltage Offset Correction: v_off_ini = 340.088 uV, v_off_fin = 323.723 uV 10:01:46 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4998 mm dXY = -0.4999 mm VT = -157.085 uVs 10:01:46 Integrated Voltage Measurement... 10:01:55 Voltage Offset Correction: v_off_ini = 337.588 uV, v_off_fin = 339.103 uV 10:01:55 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = 152.621 uVs 10:01:55 Integrated Voltage Measurement... 10:02:04 Voltage Offset Correction: v_off_ini = 335.355 uV, v_off_fin = 335.624 uV 10:02:04 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -143.405 uVs 10:02:04 Integrated Voltage Measurement... 10:02:13 Voltage Offset Correction: v_off_ini = 337.835 uV, v_off_fin = 332.095 uV 10:02:13 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 151.198 uVs 10:02:13 Integrated Voltage Measurement... 10:02:22 Voltage Offset Correction: v_off_ini = 328.221 uV, v_off_fin = 328.281 uV 10:02:22 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -163.912 uVs 10:02:22 Integrated Voltage Measurement... 10:02:31 Voltage Offset Correction: v_off_ini = 326.846 uV, v_off_fin = 339.833 uV 10:02:31 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 151.984 uVs 10:02:36 I1X Average and RMS of Measurements VT = -153.367 +- 4.471 uVs I1X = 30.676 +- 0.895 uTm 10:02:38 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:02:42 Integrated Voltage Measurement... 10:02:51 Voltage Offset Correction: v_off_ini = 332.455 uV, v_off_fin = 323.617 uV 10:02:51 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = -203.518 uVs 10:02:51 Integrated Voltage Measurement... 10:03:00 Voltage Offset Correction: v_off_ini = 333.255 uV, v_off_fin = 348.332 uV 10:03:00 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 208.803 uVs 10:03:00 Integrated Voltage Measurement... 10:03:09 Voltage Offset Correction: v_off_ini = 329.636 uV, v_off_fin = 341.047 uV 10:03:09 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -202.072 uVs 10:03:09 Integrated Voltage Measurement... 10:03:18 Voltage Offset Correction: v_off_ini = 338.349 uV, v_off_fin = 363.445 uV 10:03:18 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 207.146 uVs 10:03:18 Integrated Voltage Measurement... 10:03:27 Voltage Offset Correction: v_off_ini = 346.365 uV, v_off_fin = 336.090 uV 10:03:27 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -193.524 uVs 10:03:27 Integrated Voltage Measurement... 10:03:36 Voltage Offset Correction: v_off_ini = 340.737 uV, v_off_fin = 351.631 uV 10:03:36 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 198.257 uVs 10:03:41 I1Y Average and RMS of Measurements VT = -202.220 +- 4.520 uVs I1Y = -40.445 +- 0.905 uTm 10:03:46 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:03:48 Integrated Voltage Measurement... 10:03:57 Voltage Offset Correction: v_off_ini = 342.661 uV, v_off_fin = 332.635 uV 10:03:57 Integrated Voltage Measurement dXYus = -1.0002 mm, dXYds = 0.0000 mm dXY = -1.0002 mm VT = -134.106 uVs 10:03:57 Integrated Voltage Measurement... 10:04:06 Voltage Offset Correction: v_off_ini = 332.456 uV, v_off_fin = 336.795 uV 10:04:06 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 106.294 uVs 10:04:06 Integrated Voltage Measurement... 10:04:15 Voltage Offset Correction: v_off_ini = 318.961 uV, v_off_fin = 319.981 uV 10:04:15 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -119.401 uVs 10:04:15 Integrated Voltage Measurement... 10:04:24 Voltage Offset Correction: v_off_ini = 320.918 uV, v_off_fin = 339.939 uV 10:04:24 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 128.958 uVs 10:04:24 Integrated Voltage Measurement... 10:04:33 Voltage Offset Correction: v_off_ini = 330.354 uV, v_off_fin = 324.654 uV 10:04:33 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -140.487 uVs 10:04:33 Integrated Voltage Measurement... 10:04:42 Voltage Offset Correction: v_off_ini = 327.198 uV, v_off_fin = 343.090 uV 10:04:42 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 121.496 uVs 10:04:45 I2X Average and RMS of Measurements VT = -125.124 +- 4.456 uVs I2X = 11.511 +- 0.410 uTm^2 10:04:50 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:04:53 Integrated Voltage Measurement... 10:05:02 Voltage Offset Correction: v_off_ini = 323.934 uV, v_off_fin = 332.150 uV 10:05:02 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -510.692 uVs 10:05:02 Integrated Voltage Measurement... 10:05:11 Voltage Offset Correction: v_off_ini = 340.402 uV, v_off_fin = 334.119 uV 10:05:11 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 519.702 uVs 10:05:11 Integrated Voltage Measurement... 10:05:20 Voltage Offset Correction: v_off_ini = 326.373 uV, v_off_fin = 323.638 uV 10:05:20 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -492.301 uVs 10:05:20 Integrated Voltage Measurement... 10:05:29 Voltage Offset Correction: v_off_ini = 323.034 uV, v_off_fin = 335.505 uV 10:05:29 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 511.121 uVs 10:05:29 Integrated Voltage Measurement... 10:05:38 Voltage Offset Correction: v_off_ini = 333.773 uV, v_off_fin = 338.602 uV 10:05:38 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -505.315 uVs 10:05:38 Integrated Voltage Measurement... 10:05:47 Voltage Offset Correction: v_off_ini = 343.514 uV, v_off_fin = 336.272 uV 10:05:47 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 498.221 uVs 10:05:49 I2Y Average and RMS of Measurements VT = -506.225 +- 6.344 uVs I2Y = -46.573 +- 0.584 uTm^2 10:06:02 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:06:06 Integrated Voltage Measurement... 10:06:15 Voltage Offset Correction: v_off_ini = 338.314 uV, v_off_fin = 329.985 uV 10:06:15 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4998 mm dXY = -0.4999 mm VT = -164.042 uVs 10:06:15 Integrated Voltage Measurement... 10:06:24 Voltage Offset Correction: v_off_ini = 320.912 uV, v_off_fin = 330.048 uV 10:06:24 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 158.522 uVs 10:06:24 Integrated Voltage Measurement... 10:06:33 Voltage Offset Correction: v_off_ini = 322.973 uV, v_off_fin = 338.139 uV 10:06:33 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -168.612 uVs 10:06:33 Integrated Voltage Measurement... 10:06:42 Voltage Offset Correction: v_off_ini = 338.693 uV, v_off_fin = 335.745 uV 10:06:42 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 167.059 uVs 10:06:43 Integrated Voltage Measurement... 10:06:51 Voltage Offset Correction: v_off_ini = 338.461 uV, v_off_fin = 338.687 uV 10:06:52 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -165.448 uVs 10:06:52 Integrated Voltage Measurement... 10:07:00 Voltage Offset Correction: v_off_ini = 334.905 uV, v_off_fin = 341.471 uV 10:07:01 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 163.363 uVs 10:07:05 I1X Average and RMS of Measurements VT = -164.508 +- 2.676 uVs I1X = 32.903 +- 0.535 uTm 10:07:07 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:07:12 Integrated Voltage Measurement... 10:07:20 Voltage Offset Correction: v_off_ini = 328.135 uV, v_off_fin = 333.304 uV 10:07:21 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = -184.199 uVs 10:07:21 Integrated Voltage Measurement... 10:07:29 Voltage Offset Correction: v_off_ini = 337.457 uV, v_off_fin = 352.644 uV 10:07:29 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 211.650 uVs 10:07:30 Integrated Voltage Measurement... 10:07:38 Voltage Offset Correction: v_off_ini = 334.322 uV, v_off_fin = 328.098 uV 10:07:38 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -201.389 uVs 10:07:39 Integrated Voltage Measurement... 10:07:47 Voltage Offset Correction: v_off_ini = 336.311 uV, v_off_fin = 353.175 uV 10:07:47 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 202.465 uVs 10:07:48 Integrated Voltage Measurement... 10:07:56 Voltage Offset Correction: v_off_ini = 349.419 uV, v_off_fin = 338.416 uV 10:07:56 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -193.651 uVs 10:07:56 Integrated Voltage Measurement... 10:08:05 Voltage Offset Correction: v_off_ini = 358.969 uV, v_off_fin = 374.478 uV 10:08:05 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 209.348 uVs 10:08:10 I1Y Average and RMS of Measurements VT = -200.450 +- 1.794 uVs I1Y = -40.091 +- 0.358 uTm 10:08:15 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:08:17 Integrated Voltage Measurement... 10:08:26 Voltage Offset Correction: v_off_ini = 341.545 uV, v_off_fin = 333.487 uV 10:08:26 Integrated Voltage Measurement dXYus = -1.0001 mm, dXYds = 0.0000 mm dXY = -1.0001 mm VT = -124.547 uVs 10:08:26 Integrated Voltage Measurement... 10:08:35 Voltage Offset Correction: v_off_ini = 337.887 uV, v_off_fin = 347.383 uV 10:08:35 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 124.645 uVs 10:08:35 Integrated Voltage Measurement... 10:08:44 Voltage Offset Correction: v_off_ini = 327.614 uV, v_off_fin = 327.015 uV 10:08:44 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -135.899 uVs 10:08:45 Integrated Voltage Measurement... 10:08:53 Voltage Offset Correction: v_off_ini = 335.851 uV, v_off_fin = 348.612 uV 10:08:54 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 147.753 uVs 10:08:54 Integrated Voltage Measurement... 10:09:03 Voltage Offset Correction: v_off_ini = 346.946 uV, v_off_fin = 338.282 uV 10:09:03 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -131.139 uVs 10:09:03 Integrated Voltage Measurement... 10:09:12 Voltage Offset Correction: v_off_ini = 335.574 uV, v_off_fin = 351.746 uV 10:09:12 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 147.048 uVs 10:09:14 I2X Average and RMS of Measurements VT = -135.172 +- 7.561 uVs I2X = 12.436 +- 0.696 uTm^2 10:09:19 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:09:22 Integrated Voltage Measurement... 10:09:31 Voltage Offset Correction: v_off_ini = 333.812 uV, v_off_fin = 324.263 uV 10:09:31 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -515.441 uVs 10:09:31 Integrated Voltage Measurement... 10:09:40 Voltage Offset Correction: v_off_ini = 333.697 uV, v_off_fin = 333.143 uV 10:09:40 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 493.930 uVs 10:09:40 Integrated Voltage Measurement... 10:09:49 Voltage Offset Correction: v_off_ini = 336.519 uV, v_off_fin = 333.588 uV 10:09:49 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -502.724 uVs 10:09:49 Integrated Voltage Measurement... 10:09:58 Voltage Offset Correction: v_off_ini = 327.042 uV, v_off_fin = 342.389 uV 10:09:58 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 503.609 uVs 10:09:58 Integrated Voltage Measurement... 10:10:07 Voltage Offset Correction: v_off_ini = 331.315 uV, v_off_fin = 326.231 uV 10:10:07 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -505.013 uVs 10:10:07 Integrated Voltage Measurement... 10:10:16 Voltage Offset Correction: v_off_ini = 323.617 uV, v_off_fin = 319.595 uV 10:10:16 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 513.488 uVs 10:10:18 I2Y Average and RMS of Measurements VT = -505.701 +- 2.586 uVs I2Y = -46.524 +- 0.238 uTm^2 10:10:31 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:10:36 Integrated Voltage Measurement... 10:10:45 Voltage Offset Correction: v_off_ini = 353.894 uV, v_off_fin = 352.619 uV 10:10:45 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -178.269 uVs 10:10:45 Integrated Voltage Measurement... 10:10:54 Voltage Offset Correction: v_off_ini = 344.134 uV, v_off_fin = 347.863 uV 10:10:54 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 171.996 uVs 10:10:54 Integrated Voltage Measurement... 10:11:03 Voltage Offset Correction: v_off_ini = 347.462 uV, v_off_fin = 350.481 uV 10:11:03 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -138.136 uVs 10:11:03 Integrated Voltage Measurement... 10:11:12 Voltage Offset Correction: v_off_ini = 356.831 uV, v_off_fin = 355.247 uV 10:11:12 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 159.438 uVs 10:11:12 Integrated Voltage Measurement... 10:11:21 Voltage Offset Correction: v_off_ini = 351.113 uV, v_off_fin = 351.606 uV 10:11:21 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -169.722 uVs 10:11:21 Integrated Voltage Measurement... 10:11:30 Voltage Offset Correction: v_off_ini = 352.538 uV, v_off_fin = 360.153 uV 10:11:30 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 138.554 uVs 10:11:34 I1X Average and RMS of Measurements VT = -159.352 +- 11.370 uVs I1X = 31.871 +- 2.274 uTm 10:11:36 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:11:41 Integrated Voltage Measurement... 10:11:50 Voltage Offset Correction: v_off_ini = 338.680 uV, v_off_fin = 341.346 uV 10:11:50 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = -188.263 uVs 10:11:50 Integrated Voltage Measurement... 10:11:59 Voltage Offset Correction: v_off_ini = 352.821 uV, v_off_fin = 370.971 uV 10:11:59 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 201.401 uVs 10:11:59 Integrated Voltage Measurement... 10:12:08 Voltage Offset Correction: v_off_ini = 353.956 uV, v_off_fin = 344.835 uV 10:12:08 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -205.588 uVs 10:12:08 Integrated Voltage Measurement... 10:12:17 Voltage Offset Correction: v_off_ini = 357.326 uV, v_off_fin = 371.466 uV 10:12:17 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 205.203 uVs 10:12:17 Integrated Voltage Measurement... 10:12:26 Voltage Offset Correction: v_off_ini = 353.401 uV, v_off_fin = 354.150 uV 10:12:26 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -191.851 uVs 10:12:26 Integrated Voltage Measurement... 10:12:35 Voltage Offset Correction: v_off_ini = 359.429 uV, v_off_fin = 368.762 uV 10:12:35 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 200.284 uVs 10:12:39 I1Y Average and RMS of Measurements VT = -198.765 +- 4.716 uVs I1Y = -39.753 +- 0.942 uTm 10:12:44 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:12:47 Integrated Voltage Measurement... 10:12:56 Voltage Offset Correction: v_off_ini = 360.338 uV, v_off_fin = 367.432 uV 10:12:56 Integrated Voltage Measurement dXYus = -1.0002 mm, dXYds = 0.0000 mm dXY = -1.0002 mm VT = -150.456 uVs 10:12:56 Integrated Voltage Measurement... 10:13:05 Voltage Offset Correction: v_off_ini = 365.903 uV, v_off_fin = 385.703 uV 10:13:05 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 141.507 uVs 10:13:05 Integrated Voltage Measurement... 10:13:14 Voltage Offset Correction: v_off_ini = 365.767 uV, v_off_fin = 354.028 uV 10:13:14 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -140.020 uVs 10:13:14 Integrated Voltage Measurement... 10:13:23 Voltage Offset Correction: v_off_ini = 362.523 uV, v_off_fin = 381.077 uV 10:13:23 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 126.613 uVs 10:13:23 Integrated Voltage Measurement... 10:13:32 Voltage Offset Correction: v_off_ini = 385.311 uV, v_off_fin = 379.814 uV 10:13:32 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -153.959 uVs 10:13:32 Integrated Voltage Measurement... 10:13:41 Voltage Offset Correction: v_off_ini = 379.811 uV, v_off_fin = 389.099 uV 10:13:41 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 158.773 uVs 10:13:43 I2X Average and RMS of Measurements VT = -145.221 +- 9.425 uVs I2X = 13.360 +- 0.867 uTm^2 10:13:48 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:13:51 Integrated Voltage Measurement... 10:14:00 Voltage Offset Correction: v_off_ini = 389.523 uV, v_off_fin = 375.271 uV 10:14:00 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -496.423 uVs 10:14:00 Integrated Voltage Measurement... 10:14:09 Voltage Offset Correction: v_off_ini = 379.365 uV, v_off_fin = 371.839 uV 10:14:09 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 488.303 uVs 10:14:09 Integrated Voltage Measurement... 10:14:18 Voltage Offset Correction: v_off_ini = 402.835 uV, v_off_fin = 396.010 uV 10:14:18 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -497.316 uVs 10:14:18 Integrated Voltage Measurement... 10:14:27 Voltage Offset Correction: v_off_ini = 397.170 uV, v_off_fin = 398.018 uV 10:14:27 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 481.601 uVs 10:14:27 Integrated Voltage Measurement... 10:14:36 Voltage Offset Correction: v_off_ini = 394.667 uV, v_off_fin = 392.863 uV 10:14:36 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -493.241 uVs 10:14:36 Integrated Voltage Measurement... 10:14:45 Voltage Offset Correction: v_off_ini = 394.341 uV, v_off_fin = 393.378 uV 10:14:45 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 487.114 uVs 10:14:48 I2Y Average and RMS of Measurements VT = -490.666 +- 1.235 uVs I2Y = -45.141 +- 0.114 uTm^2 10:15:00 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:15:05 Integrated Voltage Measurement... 10:15:14 Voltage Offset Correction: v_off_ini = 394.606 uV, v_off_fin = 385.515 uV 10:15:14 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -167.279 uVs 10:15:14 Integrated Voltage Measurement... 10:15:23 Voltage Offset Correction: v_off_ini = 397.059 uV, v_off_fin = 394.319 uV 10:15:23 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 176.368 uVs 10:15:23 Integrated Voltage Measurement... 10:15:32 Voltage Offset Correction: v_off_ini = 400.457 uV, v_off_fin = 397.129 uV 10:15:32 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -173.098 uVs 10:15:32 Integrated Voltage Measurement... 10:15:41 Voltage Offset Correction: v_off_ini = 396.193 uV, v_off_fin = 402.867 uV 10:15:41 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 165.580 uVs 10:15:41 Integrated Voltage Measurement... 10:15:50 Voltage Offset Correction: v_off_ini = 400.406 uV, v_off_fin = 401.865 uV 10:15:50 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -153.463 uVs 10:15:50 Integrated Voltage Measurement... 10:15:59 Voltage Offset Correction: v_off_ini = 406.263 uV, v_off_fin = 404.843 uV 10:15:59 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 174.847 uVs 10:16:03 I1X Average and RMS of Measurements VT = -168.439 +- 3.194 uVs I1X = 33.688 +- 0.639 uTm 10:16:06 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:16:10 Integrated Voltage Measurement... 10:16:19 Voltage Offset Correction: v_off_ini = 397.248 uV, v_off_fin = 379.563 uV 10:16:19 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = -191.698 uVs 10:16:19 Integrated Voltage Measurement... 10:16:28 Voltage Offset Correction: v_off_ini = 402.595 uV, v_off_fin = 414.810 uV 10:16:28 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 193.235 uVs 10:16:28 Integrated Voltage Measurement... 10:16:37 Voltage Offset Correction: v_off_ini = 407.358 uV, v_off_fin = 387.564 uV 10:16:37 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -213.628 uVs 10:16:37 Integrated Voltage Measurement... 10:16:46 Voltage Offset Correction: v_off_ini = 396.640 uV, v_off_fin = 406.083 uV 10:16:46 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 200.217 uVs 10:16:46 Integrated Voltage Measurement... 10:16:55 Voltage Offset Correction: v_off_ini = 402.938 uV, v_off_fin = 389.469 uV 10:16:55 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -201.070 uVs 10:16:55 Integrated Voltage Measurement... 10:17:04 Voltage Offset Correction: v_off_ini = 397.132 uV, v_off_fin = 409.416 uV 10:17:04 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 197.151 uVs 10:17:08 I1Y Average and RMS of Measurements VT = -199.500 +- 5.908 uVs I1Y = -39.899 +- 1.180 uTm 10:17:13 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:17:16 Integrated Voltage Measurement... 10:17:25 Voltage Offset Correction: v_off_ini = 393.022 uV, v_off_fin = 389.002 uV 10:17:25 Integrated Voltage Measurement dXYus = -1.0002 mm, dXYds = 0.0000 mm dXY = -1.0002 mm VT = -138.936 uVs 10:17:25 Integrated Voltage Measurement... 10:17:34 Voltage Offset Correction: v_off_ini = 400.026 uV, v_off_fin = 412.742 uV 10:17:34 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 151.370 uVs 10:17:34 Integrated Voltage Measurement... 10:17:43 Voltage Offset Correction: v_off_ini = 408.162 uV, v_off_fin = 402.951 uV 10:17:43 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -161.983 uVs 10:17:43 Integrated Voltage Measurement... 10:17:52 Voltage Offset Correction: v_off_ini = 417.898 uV, v_off_fin = 411.596 uV 10:17:52 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 134.694 uVs 10:17:52 Integrated Voltage Measurement... 10:18:01 Voltage Offset Correction: v_off_ini = 405.301 uV, v_off_fin = 399.249 uV 10:18:01 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -139.979 uVs 10:18:01 Integrated Voltage Measurement... 10:18:10 Voltage Offset Correction: v_off_ini = 415.349 uV, v_off_fin = 418.725 uV 10:18:10 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 138.789 uVs 10:18:12 I2X Average and RMS of Measurements VT = -144.292 +- 3.706 uVs I2X = 13.274 +- 0.341 uTm^2 10:18:18 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:18:20 Integrated Voltage Measurement... 10:18:29 Voltage Offset Correction: v_off_ini = 402.212 uV, v_off_fin = 403.597 uV 10:18:29 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -494.788 uVs 10:18:29 Integrated Voltage Measurement... 10:18:38 Voltage Offset Correction: v_off_ini = 415.933 uV, v_off_fin = 411.484 uV 10:18:38 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 493.728 uVs 10:18:38 Integrated Voltage Measurement... 10:18:47 Voltage Offset Correction: v_off_ini = 411.759 uV, v_off_fin = 412.366 uV 10:18:47 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -467.569 uVs 10:18:47 Integrated Voltage Measurement... 10:18:56 Voltage Offset Correction: v_off_ini = 423.746 uV, v_off_fin = 419.096 uV 10:18:56 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 482.983 uVs 10:18:56 Integrated Voltage Measurement... 10:19:05 Voltage Offset Correction: v_off_ini = 410.620 uV, v_off_fin = 405.504 uV 10:19:05 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -474.109 uVs 10:19:05 Integrated Voltage Measurement... 10:19:14 Voltage Offset Correction: v_off_ini = 407.374 uV, v_off_fin = 410.348 uV 10:19:14 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 487.213 uVs 10:19:17 I2Y Average and RMS of Measurements VT = -483.398 +- 7.987 uVs I2Y = -44.473 +- 0.735 uTm^2 10:19:30 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:19:34 Integrated Voltage Measurement... 10:19:43 Voltage Offset Correction: v_off_ini = 416.330 uV, v_off_fin = 398.423 uV 10:19:43 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -215.215 uVs 10:19:43 Integrated Voltage Measurement... 10:19:52 Voltage Offset Correction: v_off_ini = 399.387 uV, v_off_fin = 399.838 uV 10:19:52 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 179.397 uVs 10:19:52 Integrated Voltage Measurement... 10:20:01 Voltage Offset Correction: v_off_ini = 410.564 uV, v_off_fin = 408.577 uV 10:20:01 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -164.791 uVs 10:20:01 Integrated Voltage Measurement... 10:20:10 Voltage Offset Correction: v_off_ini = 409.209 uV, v_off_fin = 412.784 uV 10:20:10 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 178.857 uVs 10:20:10 Integrated Voltage Measurement... 10:20:19 Voltage Offset Correction: v_off_ini = 428.300 uV, v_off_fin = 402.421 uV 10:20:19 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -174.608 uVs 10:20:19 Integrated Voltage Measurement... 10:20:28 Voltage Offset Correction: v_off_ini = 410.240 uV, v_off_fin = 414.686 uV 10:20:28 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 160.464 uVs 10:20:32 I1X Average and RMS of Measurements VT = -178.889 +- 13.140 uVs I1X = 35.778 +- 2.628 uTm 10:20:35 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:20:39 Integrated Voltage Measurement... 10:20:48 Voltage Offset Correction: v_off_ini = 412.747 uV, v_off_fin = 399.077 uV 10:20:48 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4998 mm dXY = 0.4999 mm VT = -193.858 uVs 10:20:48 Integrated Voltage Measurement... 10:20:57 Voltage Offset Correction: v_off_ini = 408.018 uV, v_off_fin = 416.820 uV 10:20:57 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 199.681 uVs 10:20:57 Integrated Voltage Measurement... 10:21:06 Voltage Offset Correction: v_off_ini = 411.056 uV, v_off_fin = 396.882 uV 10:21:06 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -212.851 uVs 10:21:06 Integrated Voltage Measurement... 10:21:15 Voltage Offset Correction: v_off_ini = 416.756 uV, v_off_fin = 424.531 uV 10:21:15 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 198.709 uVs 10:21:15 Integrated Voltage Measurement... 10:21:24 Voltage Offset Correction: v_off_ini = 417.100 uV, v_off_fin = 405.375 uV 10:21:24 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -194.143 uVs 10:21:24 Integrated Voltage Measurement... 10:21:33 Voltage Offset Correction: v_off_ini = 426.706 uV, v_off_fin = 442.901 uV 10:21:33 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 185.095 uVs 10:21:37 I1Y Average and RMS of Measurements VT = -197.390 +- 6.612 uVs I1Y = -39.476 +- 1.322 uTm 10:21:42 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:21:45 Integrated Voltage Measurement... 10:21:54 Voltage Offset Correction: v_off_ini = 435.064 uV, v_off_fin = 427.094 uV 10:21:54 Integrated Voltage Measurement dXYus = -1.0002 mm, dXYds = 0.0000 mm dXY = -1.0002 mm VT = -189.334 uVs 10:21:54 Integrated Voltage Measurement... 10:22:03 Voltage Offset Correction: v_off_ini = 439.556 uV, v_off_fin = 441.266 uV 10:22:03 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 164.036 uVs 10:22:03 Integrated Voltage Measurement... 10:22:12 Voltage Offset Correction: v_off_ini = 431.057 uV, v_off_fin = 425.687 uV 10:22:12 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -144.776 uVs 10:22:12 Integrated Voltage Measurement... 10:22:21 Voltage Offset Correction: v_off_ini = 434.645 uV, v_off_fin = 448.223 uV 10:22:21 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 146.472 uVs 10:22:21 Integrated Voltage Measurement... 10:22:30 Voltage Offset Correction: v_off_ini = 436.144 uV, v_off_fin = 426.475 uV 10:22:30 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -152.529 uVs 10:22:30 Integrated Voltage Measurement... 10:22:39 Voltage Offset Correction: v_off_ini = 453.388 uV, v_off_fin = 451.717 uV 10:22:39 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 142.129 uVs 10:22:41 I2X Average and RMS of Measurements VT = -156.546 +- 14.258 uVs I2X = 14.402 +- 1.311 uTm^2 10:22:46 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:22:49 Integrated Voltage Measurement... 10:22:58 Voltage Offset Correction: v_off_ini = 435.315 uV, v_off_fin = 441.804 uV 10:22:58 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -474.778 uVs 10:22:58 Integrated Voltage Measurement... 10:23:07 Voltage Offset Correction: v_off_ini = 443.177 uV, v_off_fin = 448.102 uV 10:23:07 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 485.933 uVs 10:23:07 Integrated Voltage Measurement... 10:23:16 Voltage Offset Correction: v_off_ini = 452.308 uV, v_off_fin = 448.971 uV 10:23:16 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -477.919 uVs 10:23:16 Integrated Voltage Measurement... 10:23:25 Voltage Offset Correction: v_off_ini = 457.489 uV, v_off_fin = 453.094 uV 10:23:25 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 485.903 uVs 10:23:25 Integrated Voltage Measurement... 10:23:34 Voltage Offset Correction: v_off_ini = 455.046 uV, v_off_fin = 453.002 uV 10:23:34 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -486.846 uVs 10:23:34 Integrated Voltage Measurement... 10:23:43 Voltage Offset Correction: v_off_ini = 450.104 uV, v_off_fin = 459.798 uV 10:23:43 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 477.166 uVs 10:23:45 I2Y Average and RMS of Measurements VT = -481.424 +- 0.757 uVs I2Y = -44.291 +- 0.070 uTm^2 10:23:57 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:24:02 Integrated Voltage Measurement... 10:24:10 Voltage Offset Correction: v_off_ini = 466.628 uV, v_off_fin = 456.359 uV 10:24:10 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -218.622 uVs 10:24:11 Integrated Voltage Measurement... 10:24:19 Voltage Offset Correction: v_off_ini = 433.906 uV, v_off_fin = 451.139 uV 10:24:19 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 189.284 uVs 10:24:20 Integrated Voltage Measurement... 10:24:28 Voltage Offset Correction: v_off_ini = 444.244 uV, v_off_fin = 453.854 uV 10:24:29 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -176.256 uVs 10:24:29 Integrated Voltage Measurement... 10:24:37 Voltage Offset Correction: v_off_ini = 443.236 uV, v_off_fin = 452.001 uV 10:24:38 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 172.989 uVs 10:24:38 Integrated Voltage Measurement... 10:24:46 Voltage Offset Correction: v_off_ini = 449.888 uV, v_off_fin = 447.542 uV 10:24:47 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -178.817 uVs 10:24:47 Integrated Voltage Measurement... 10:24:55 Voltage Offset Correction: v_off_ini = 435.114 uV, v_off_fin = 449.408 uV 10:24:56 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 182.744 uVs 10:25:00 I1X Average and RMS of Measurements VT = -186.452 +- 12.628 uVs I1X = 37.292 +- 2.525 uTm 10:25:02 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:25:06 Integrated Voltage Measurement... 10:25:15 Voltage Offset Correction: v_off_ini = 450.578 uV, v_off_fin = 447.608 uV 10:25:15 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = -198.165 uVs 10:25:15 Integrated Voltage Measurement... 10:25:24 Voltage Offset Correction: v_off_ini = 452.475 uV, v_off_fin = 460.030 uV 10:25:24 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 191.004 uVs 10:25:24 Integrated Voltage Measurement... 10:25:33 Voltage Offset Correction: v_off_ini = 452.632 uV, v_off_fin = 436.788 uV 10:25:33 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -205.353 uVs 10:25:33 Integrated Voltage Measurement... 10:25:42 Voltage Offset Correction: v_off_ini = 442.384 uV, v_off_fin = 467.657 uV 10:25:42 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 192.487 uVs 10:25:42 Integrated Voltage Measurement... 10:25:51 Voltage Offset Correction: v_off_ini = 458.705 uV, v_off_fin = 439.841 uV 10:25:51 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -225.095 uVs 10:25:51 Integrated Voltage Measurement... 10:26:00 Voltage Offset Correction: v_off_ini = 450.006 uV, v_off_fin = 453.303 uV 10:26:00 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 187.672 uVs 10:26:05 I1Y Average and RMS of Measurements VT = -199.963 +- 4.873 uVs I1Y = -39.990 +- 0.974 uTm 10:26:10 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:26:12 Integrated Voltage Measurement... 10:26:21 Voltage Offset Correction: v_off_ini = 441.089 uV, v_off_fin = 434.350 uV 10:26:21 Integrated Voltage Measurement dXYus = -1.0002 mm, dXYds = 0.0000 mm dXY = -1.0002 mm VT = -176.298 uVs 10:26:21 Integrated Voltage Measurement... 10:26:30 Voltage Offset Correction: v_off_ini = 436.067 uV, v_off_fin = 446.010 uV 10:26:30 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 158.802 uVs 10:26:30 Integrated Voltage Measurement... 10:26:39 Voltage Offset Correction: v_off_ini = 443.189 uV, v_off_fin = 431.304 uV 10:26:39 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -165.351 uVs 10:26:39 Integrated Voltage Measurement... 10:26:48 Voltage Offset Correction: v_off_ini = 431.963 uV, v_off_fin = 458.292 uV 10:26:48 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 143.900 uVs 10:26:48 Integrated Voltage Measurement... 10:26:57 Voltage Offset Correction: v_off_ini = 435.529 uV, v_off_fin = 428.489 uV 10:26:57 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -154.681 uVs 10:26:57 Integrated Voltage Measurement... 10:27:06 Voltage Offset Correction: v_off_ini = 442.207 uV, v_off_fin = 459.607 uV 10:27:06 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 140.725 uVs 10:27:09 I2X Average and RMS of Measurements VT = -156.626 +- 8.225 uVs I2X = 14.410 +- 0.756 uTm^2 10:27:14 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:27:16 Integrated Voltage Measurement... 10:27:25 Voltage Offset Correction: v_off_ini = 446.103 uV, v_off_fin = 437.152 uV 10:27:25 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -475.673 uVs 10:27:25 Integrated Voltage Measurement... 10:27:34 Voltage Offset Correction: v_off_ini = 440.446 uV, v_off_fin = 447.888 uV 10:27:34 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 484.534 uVs 10:27:35 Integrated Voltage Measurement... 10:27:43 Voltage Offset Correction: v_off_ini = 453.397 uV, v_off_fin = 443.394 uV 10:27:43 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -480.774 uVs 10:27:44 Integrated Voltage Measurement... 10:27:52 Voltage Offset Correction: v_off_ini = 453.281 uV, v_off_fin = 446.983 uV 10:27:53 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 478.299 uVs 10:27:53 Integrated Voltage Measurement... 10:28:02 Voltage Offset Correction: v_off_ini = 448.021 uV, v_off_fin = 450.094 uV 10:28:02 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -472.231 uVs 10:28:02 Integrated Voltage Measurement... 10:28:11 Voltage Offset Correction: v_off_ini = 457.714 uV, v_off_fin = 437.284 uV 10:28:11 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 483.177 uVs 10:28:13 I2Y Average and RMS of Measurements VT = -479.115 +- 1.024 uVs I2Y = -44.079 +- 0.094 uTm^2 10:28:25 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:28:29 Integrated Voltage Measurement... 10:28:38 Voltage Offset Correction: v_off_ini = 456.178 uV, v_off_fin = 454.535 uV 10:28:38 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -160.809 uVs 10:28:38 Integrated Voltage Measurement... 10:28:47 Voltage Offset Correction: v_off_ini = 457.118 uV, v_off_fin = 461.985 uV 10:28:47 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 173.482 uVs 10:28:47 Integrated Voltage Measurement... 10:28:56 Voltage Offset Correction: v_off_ini = 458.009 uV, v_off_fin = 445.499 uV 10:28:56 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -168.624 uVs 10:28:56 Integrated Voltage Measurement... 10:29:05 Voltage Offset Correction: v_off_ini = 460.605 uV, v_off_fin = 463.970 uV 10:29:05 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 171.956 uVs 10:29:05 Integrated Voltage Measurement... 10:29:14 Voltage Offset Correction: v_off_ini = 468.460 uV, v_off_fin = 450.321 uV 10:29:14 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -177.577 uVs 10:29:14 Integrated Voltage Measurement... 10:29:23 Voltage Offset Correction: v_off_ini = 458.411 uV, v_off_fin = 459.253 uV 10:29:23 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 169.774 uVs 10:29:27 I1X Average and RMS of Measurements VT = -170.370 +- 2.666 uVs I1X = 34.074 +- 0.533 uTm 10:29:30 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:29:34 Integrated Voltage Measurement... 10:29:43 Voltage Offset Correction: v_off_ini = 443.682 uV, v_off_fin = 437.490 uV 10:29:43 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = -195.550 uVs 10:29:43 Integrated Voltage Measurement... 10:29:52 Voltage Offset Correction: v_off_ini = 460.506 uV, v_off_fin = 478.292 uV 10:29:52 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 201.224 uVs 10:29:52 Integrated Voltage Measurement... 10:30:01 Voltage Offset Correction: v_off_ini = 464.824 uV, v_off_fin = 452.016 uV 10:30:01 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -210.886 uVs 10:30:01 Integrated Voltage Measurement... 10:30:10 Voltage Offset Correction: v_off_ini = 456.214 uV, v_off_fin = 474.707 uV 10:30:10 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 189.442 uVs 10:30:10 Integrated Voltage Measurement... 10:30:19 Voltage Offset Correction: v_off_ini = 460.857 uV, v_off_fin = 454.306 uV 10:30:19 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -205.777 uVs 10:30:19 Integrated Voltage Measurement... 10:30:28 Voltage Offset Correction: v_off_ini = 469.839 uV, v_off_fin = 477.668 uV 10:30:28 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 179.916 uVs 10:30:32 I1Y Average and RMS of Measurements VT = -197.133 +- 3.116 uVs I1Y = -39.424 +- 0.624 uTm 10:30:37 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:30:40 Integrated Voltage Measurement... 10:30:49 Voltage Offset Correction: v_off_ini = 471.538 uV, v_off_fin = 456.924 uV 10:30:49 Integrated Voltage Measurement dXYus = -1.0002 mm, dXYds = 0.0000 mm dXY = -1.0002 mm VT = -167.885 uVs 10:30:49 Integrated Voltage Measurement... 10:30:58 Voltage Offset Correction: v_off_ini = 478.811 uV, v_off_fin = 479.558 uV 10:30:58 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 153.995 uVs 10:30:58 Integrated Voltage Measurement... 10:31:07 Voltage Offset Correction: v_off_ini = 468.896 uV, v_off_fin = 464.872 uV 10:31:07 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -160.959 uVs 10:31:07 Integrated Voltage Measurement... 10:31:16 Voltage Offset Correction: v_off_ini = 468.803 uV, v_off_fin = 476.443 uV 10:31:16 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 165.602 uVs 10:31:16 Integrated Voltage Measurement... 10:31:25 Voltage Offset Correction: v_off_ini = 471.936 uV, v_off_fin = 458.336 uV 10:31:25 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -159.975 uVs 10:31:25 Integrated Voltage Measurement... 10:31:34 Voltage Offset Correction: v_off_ini = 466.669 uV, v_off_fin = 476.724 uV 10:31:34 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 178.281 uVs 10:31:36 I2X Average and RMS of Measurements VT = -164.450 +- 3.443 uVs I2X = 15.129 +- 0.318 uTm^2 10:31:41 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:31:44 Integrated Voltage Measurement... 10:31:53 Voltage Offset Correction: v_off_ini = 459.137 uV, v_off_fin = 462.359 uV 10:31:53 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -465.705 uVs 10:31:53 Integrated Voltage Measurement... 10:32:02 Voltage Offset Correction: v_off_ini = 451.884 uV, v_off_fin = 458.201 uV 10:32:02 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 463.854 uVs 10:32:02 Integrated Voltage Measurement... 10:32:11 Voltage Offset Correction: v_off_ini = 466.381 uV, v_off_fin = 458.516 uV 10:32:11 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -464.957 uVs 10:32:11 Integrated Voltage Measurement... 10:32:20 Voltage Offset Correction: v_off_ini = 458.577 uV, v_off_fin = 469.487 uV 10:32:20 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 478.928 uVs 10:32:20 Integrated Voltage Measurement... 10:32:29 Voltage Offset Correction: v_off_ini = 463.329 uV, v_off_fin = 474.302 uV 10:32:29 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -462.984 uVs 10:32:29 Integrated Voltage Measurement... 10:32:38 Voltage Offset Correction: v_off_ini = 477.908 uV, v_off_fin = 486.639 uV 10:32:38 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 464.880 uVs 10:32:40 I2Y Average and RMS of Measurements VT = -466.885 +- 3.593 uVs I2Y = -42.953 +- 0.331 uTm^2 10:32:59 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:33:03 Integrated Voltage Measurement... 10:33:12 Voltage Offset Correction: v_off_ini = 491.960 uV, v_off_fin = 484.514 uV 10:33:12 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -188.261 uVs 10:33:12 Integrated Voltage Measurement... 10:33:21 Voltage Offset Correction: v_off_ini = 492.445 uV, v_off_fin = 493.668 uV 10:33:21 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 169.815 uVs 10:33:21 Integrated Voltage Measurement... 10:33:30 Voltage Offset Correction: v_off_ini = 486.809 uV, v_off_fin = 490.955 uV 10:33:30 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -175.361 uVs 10:33:30 Integrated Voltage Measurement... 10:33:39 Voltage Offset Correction: v_off_ini = 500.472 uV, v_off_fin = 504.837 uV 10:33:39 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 183.747 uVs 10:33:39 Integrated Voltage Measurement... 10:33:48 Voltage Offset Correction: v_off_ini = 499.560 uV, v_off_fin = 502.163 uV 10:33:48 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -188.365 uVs 10:33:48 Integrated Voltage Measurement... 10:33:57 Voltage Offset Correction: v_off_ini = 499.130 uV, v_off_fin = 501.450 uV 10:33:57 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 183.441 uVs 10:34:02 I1X Average and RMS of Measurements VT = -181.498 +- 3.122 uVs I1X = 36.301 +- 0.624 uTm 10:34:04 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:34:08 Integrated Voltage Measurement... 10:34:17 Voltage Offset Correction: v_off_ini = 485.184 uV, v_off_fin = 476.369 uV 10:34:17 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = -195.769 uVs 10:34:17 Integrated Voltage Measurement... 10:34:26 Voltage Offset Correction: v_off_ini = 494.393 uV, v_off_fin = 512.584 uV 10:34:26 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 191.450 uVs 10:34:26 Integrated Voltage Measurement... 10:34:35 Voltage Offset Correction: v_off_ini = 497.024 uV, v_off_fin = 484.809 uV 10:34:35 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -219.726 uVs 10:34:35 Integrated Voltage Measurement... 10:34:44 Voltage Offset Correction: v_off_ini = 493.780 uV, v_off_fin = 511.591 uV 10:34:44 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 196.666 uVs 10:34:44 Integrated Voltage Measurement... 10:34:53 Voltage Offset Correction: v_off_ini = 493.843 uV, v_off_fin = 483.488 uV 10:34:53 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -201.957 uVs 10:34:53 Integrated Voltage Measurement... 10:35:02 Voltage Offset Correction: v_off_ini = 493.470 uV, v_off_fin = 504.571 uV 10:35:02 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 191.981 uVs 10:35:06 I1Y Average and RMS of Measurements VT = -199.592 +- 6.237 uVs I1Y = -39.919 +- 1.247 uTm 10:35:11 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:35:14 Integrated Voltage Measurement... 10:35:23 Voltage Offset Correction: v_off_ini = 498.699 uV, v_off_fin = 500.156 uV 10:35:23 Integrated Voltage Measurement dXYus = -1.0002 mm, dXYds = 0.0000 mm dXY = -1.0002 mm VT = -187.573 uVs 10:35:23 Integrated Voltage Measurement... 10:35:32 Voltage Offset Correction: v_off_ini = 502.521 uV, v_off_fin = 513.077 uV 10:35:32 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 167.264 uVs 10:35:32 Integrated Voltage Measurement... 10:35:41 Voltage Offset Correction: v_off_ini = 495.370 uV, v_off_fin = 490.094 uV 10:35:41 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -156.228 uVs 10:35:41 Integrated Voltage Measurement... 10:35:50 Voltage Offset Correction: v_off_ini = 499.546 uV, v_off_fin = 502.705 uV 10:35:50 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 146.896 uVs 10:35:50 Integrated Voltage Measurement... 10:35:59 Voltage Offset Correction: v_off_ini = 492.537 uV, v_off_fin = 493.780 uV 10:35:59 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -174.957 uVs 10:35:59 Integrated Voltage Measurement... 10:36:08 Voltage Offset Correction: v_off_ini = 493.420 uV, v_off_fin = 503.739 uV 10:36:08 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 177.413 uVs 10:36:11 I2X Average and RMS of Measurements VT = -168.389 +- 11.909 uVs I2X = 15.491 +- 1.095 uTm^2 10:36:16 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:36:18 Integrated Voltage Measurement... 10:36:27 Voltage Offset Correction: v_off_ini = 503.425 uV, v_off_fin = 505.376 uV 10:36:27 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -450.824 uVs 10:36:27 Integrated Voltage Measurement... 10:36:36 Voltage Offset Correction: v_off_ini = 510.969 uV, v_off_fin = 500.196 uV 10:36:36 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 442.100 uVs 10:36:36 Integrated Voltage Measurement... 10:36:45 Voltage Offset Correction: v_off_ini = 503.848 uV, v_off_fin = 499.999 uV 10:36:45 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -442.819 uVs 10:36:45 Integrated Voltage Measurement... 10:36:54 Voltage Offset Correction: v_off_ini = 512.713 uV, v_off_fin = 500.075 uV 10:36:54 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 442.836 uVs 10:36:54 Integrated Voltage Measurement... 10:37:03 Voltage Offset Correction: v_off_ini = 500.854 uV, v_off_fin = 496.475 uV 10:37:03 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -451.679 uVs 10:37:03 Integrated Voltage Measurement... 10:37:12 Voltage Offset Correction: v_off_ini = 500.962 uV, v_off_fin = 502.422 uV 10:37:12 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 457.013 uVs 10:37:15 I2Y Average and RMS of Measurements VT = -447.878 +- 4.808 uVs I2Y = -41.205 +- 0.442 uTm^2 10:37:33 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:37:38 Integrated Voltage Measurement... 10:37:46 Voltage Offset Correction: v_off_ini = 509.599 uV, v_off_fin = 495.360 uV 10:37:47 Integrated Voltage Measurement dXYus = -0.4998 mm, dXYds = -0.4998 mm dXY = -0.4998 mm VT = -185.623 uVs 10:37:47 Integrated Voltage Measurement... 10:37:55 Voltage Offset Correction: v_off_ini = 489.881 uV, v_off_fin = 504.859 uV 10:37:55 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 186.266 uVs 10:37:56 Integrated Voltage Measurement... 10:38:04 Voltage Offset Correction: v_off_ini = 498.984 uV, v_off_fin = 508.040 uV 10:38:05 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -204.952 uVs 10:38:05 Integrated Voltage Measurement... 10:38:13 Voltage Offset Correction: v_off_ini = 507.096 uV, v_off_fin = 495.974 uV 10:38:14 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 188.238 uVs 10:38:14 Integrated Voltage Measurement... 10:38:23 Voltage Offset Correction: v_off_ini = 491.238 uV, v_off_fin = 485.834 uV 10:38:23 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -175.407 uVs 10:38:23 Integrated Voltage Measurement... 10:38:31 Voltage Offset Correction: v_off_ini = 493.041 uV, v_off_fin = 495.857 uV 10:38:32 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 156.684 uVs 10:38:36 I1X Average and RMS of Measurements VT = -182.862 +- 12.661 uVs I1X = 36.575 +- 2.532 uTm 10:38:38 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:38:43 Integrated Voltage Measurement... 10:38:51 Voltage Offset Correction: v_off_ini = 496.469 uV, v_off_fin = 489.147 uV 10:38:51 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = -201.968 uVs 10:38:52 Integrated Voltage Measurement... 10:39:00 Voltage Offset Correction: v_off_ini = 499.012 uV, v_off_fin = 498.739 uV 10:39:00 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 202.419 uVs 10:39:01 Integrated Voltage Measurement... 10:39:09 Voltage Offset Correction: v_off_ini = 492.582 uV, v_off_fin = 484.725 uV 10:39:09 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -190.798 uVs 10:39:09 Integrated Voltage Measurement... 10:39:18 Voltage Offset Correction: v_off_ini = 505.007 uV, v_off_fin = 502.382 uV 10:39:18 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 190.214 uVs 10:39:18 Integrated Voltage Measurement... 10:39:27 Voltage Offset Correction: v_off_ini = 504.140 uV, v_off_fin = 490.849 uV 10:39:27 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -193.647 uVs 10:39:27 Integrated Voltage Measurement... 10:39:36 Voltage Offset Correction: v_off_ini = 514.696 uV, v_off_fin = 524.886 uV 10:39:36 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 192.760 uVs 10:39:41 I1Y Average and RMS of Measurements VT = -195.301 +- 4.997 uVs I1Y = -39.061 +- 1.000 uTm 10:39:46 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:39:48 Integrated Voltage Measurement... 10:39:57 Voltage Offset Correction: v_off_ini = 501.475 uV, v_off_fin = 509.398 uV 10:39:57 Integrated Voltage Measurement dXYus = -1.0002 mm, dXYds = 0.0000 mm dXY = -1.0002 mm VT = -165.480 uVs 10:39:57 Integrated Voltage Measurement... 10:40:06 Voltage Offset Correction: v_off_ini = 518.467 uV, v_off_fin = 523.481 uV 10:40:06 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 143.328 uVs 10:40:06 Integrated Voltage Measurement... 10:40:15 Voltage Offset Correction: v_off_ini = 513.602 uV, v_off_fin = 509.718 uV 10:40:15 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -174.131 uVs 10:40:15 Integrated Voltage Measurement... 10:40:24 Voltage Offset Correction: v_off_ini = 512.786 uV, v_off_fin = 513.283 uV 10:40:24 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 180.573 uVs 10:40:24 Integrated Voltage Measurement... 10:40:33 Voltage Offset Correction: v_off_ini = 510.837 uV, v_off_fin = 493.729 uV 10:40:33 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -171.687 uVs 10:40:33 Integrated Voltage Measurement... 10:40:42 Voltage Offset Correction: v_off_ini = 503.333 uV, v_off_fin = 504.949 uV 10:40:42 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 152.617 uVs 10:40:45 I2X Average and RMS of Measurements VT = -164.636 +- 9.532 uVs I2X = 15.146 +- 0.877 uTm^2 10:40:50 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:40:53 Integrated Voltage Measurement... 10:41:01 Voltage Offset Correction: v_off_ini = 495.220 uV, v_off_fin = 495.736 uV 10:41:02 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -441.188 uVs 10:41:02 Integrated Voltage Measurement... 10:41:11 Voltage Offset Correction: v_off_ini = 506.473 uV, v_off_fin = 499.730 uV 10:41:11 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 436.322 uVs 10:41:11 Integrated Voltage Measurement... 10:41:20 Voltage Offset Correction: v_off_ini = 499.607 uV, v_off_fin = 497.278 uV 10:41:20 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -429.112 uVs 10:41:20 Integrated Voltage Measurement... 10:41:29 Voltage Offset Correction: v_off_ini = 498.985 uV, v_off_fin = 507.838 uV 10:41:29 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 420.810 uVs 10:41:29 Integrated Voltage Measurement... 10:41:38 Voltage Offset Correction: v_off_ini = 494.741 uV, v_off_fin = 497.501 uV 10:41:38 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -436.640 uVs 10:41:38 Integrated Voltage Measurement... 10:41:47 Voltage Offset Correction: v_off_ini = 502.100 uV, v_off_fin = 506.573 uV 10:41:47 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 442.435 uVs 10:41:49 I2Y Average and RMS of Measurements VT = -434.418 +- 6.694 uVs I2Y = -39.966 +- 0.616 uTm^2 10:42:17 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:42:22 Integrated Voltage Measurement... 10:42:31 Voltage Offset Correction: v_off_ini = 505.215 uV, v_off_fin = 505.800 uV 10:42:31 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -197.174 uVs 10:42:31 Integrated Voltage Measurement... 10:42:40 Voltage Offset Correction: v_off_ini = 514.470 uV, v_off_fin = 522.201 uV 10:42:40 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = 180.401 uVs 10:42:40 Integrated Voltage Measurement... 10:42:49 Voltage Offset Correction: v_off_ini = 510.721 uV, v_off_fin = 511.664 uV 10:42:49 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -182.998 uVs 10:42:49 Integrated Voltage Measurement... 10:42:58 Voltage Offset Correction: v_off_ini = 512.433 uV, v_off_fin = 518.896 uV 10:42:58 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 179.684 uVs 10:42:58 Integrated Voltage Measurement... 10:43:07 Voltage Offset Correction: v_off_ini = 515.993 uV, v_off_fin = 511.297 uV 10:43:07 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -165.612 uVs 10:43:07 Integrated Voltage Measurement... 10:43:16 Voltage Offset Correction: v_off_ini = 505.952 uV, v_off_fin = 514.013 uV 10:43:16 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 168.171 uVs 10:43:20 I1X Average and RMS of Measurements VT = -179.007 +- 9.090 uVs I1X = 35.803 +- 1.818 uTm 10:43:22 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:43:27 Integrated Voltage Measurement... 10:43:36 Voltage Offset Correction: v_off_ini = 495.884 uV, v_off_fin = 502.491 uV 10:43:36 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = -202.451 uVs 10:43:36 Integrated Voltage Measurement... 10:43:45 Voltage Offset Correction: v_off_ini = 498.818 uV, v_off_fin = 512.634 uV 10:43:45 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 193.885 uVs 10:43:45 Integrated Voltage Measurement... 10:43:54 Voltage Offset Correction: v_off_ini = 504.051 uV, v_off_fin = 508.133 uV 10:43:54 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -214.068 uVs 10:43:54 Integrated Voltage Measurement... 10:44:03 Voltage Offset Correction: v_off_ini = 500.820 uV, v_off_fin = 504.721 uV 10:44:03 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 186.450 uVs 10:44:03 Integrated Voltage Measurement... 10:44:11 Voltage Offset Correction: v_off_ini = 500.770 uV, v_off_fin = 490.783 uV 10:44:12 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -210.463 uVs 10:44:12 Integrated Voltage Measurement... 10:44:20 Voltage Offset Correction: v_off_ini = 504.779 uV, v_off_fin = 520.601 uV 10:44:21 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 207.188 uVs 10:44:25 I1Y Average and RMS of Measurements VT = -202.417 +- 4.611 uVs I1Y = -40.484 +- 0.922 uTm 10:44:30 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:44:32 Integrated Voltage Measurement... 10:44:41 Voltage Offset Correction: v_off_ini = 506.560 uV, v_off_fin = 501.089 uV 10:44:41 Integrated Voltage Measurement dXYus = -1.0002 mm, dXYds = 0.0000 mm dXY = -1.0002 mm VT = -177.147 uVs 10:44:41 Integrated Voltage Measurement... 10:44:50 Voltage Offset Correction: v_off_ini = 512.960 uV, v_off_fin = 516.927 uV 10:44:50 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 171.610 uVs 10:44:50 Integrated Voltage Measurement... 10:44:59 Voltage Offset Correction: v_off_ini = 515.002 uV, v_off_fin = 502.601 uV 10:44:59 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -176.420 uVs 10:44:59 Integrated Voltage Measurement... 10:45:08 Voltage Offset Correction: v_off_ini = 505.673 uV, v_off_fin = 518.823 uV 10:45:08 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 151.832 uVs 10:45:08 Integrated Voltage Measurement... 10:45:17 Voltage Offset Correction: v_off_ini = 499.776 uV, v_off_fin = 493.709 uV 10:45:17 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -152.716 uVs 10:45:18 Integrated Voltage Measurement... 10:45:26 Voltage Offset Correction: v_off_ini = 503.778 uV, v_off_fin = 518.190 uV 10:45:26 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 146.497 uVs 10:45:29 I2X Average and RMS of Measurements VT = -162.704 +- 10.163 uVs I2X = 14.969 +- 0.934 uTm^2 10:45:34 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:45:37 Integrated Voltage Measurement... 10:45:46 Voltage Offset Correction: v_off_ini = 504.584 uV, v_off_fin = 501.680 uV 10:45:46 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -382.478 uVs 10:45:46 Integrated Voltage Measurement... 10:45:55 Voltage Offset Correction: v_off_ini = 504.909 uV, v_off_fin = 501.433 uV 10:45:55 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 391.106 uVs 10:45:55 Integrated Voltage Measurement... 10:46:04 Voltage Offset Correction: v_off_ini = 506.125 uV, v_off_fin = 505.352 uV 10:46:04 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -409.404 uVs 10:46:04 Integrated Voltage Measurement... 10:46:13 Voltage Offset Correction: v_off_ini = 500.364 uV, v_off_fin = 504.609 uV 10:46:13 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 409.179 uVs 10:46:13 Integrated Voltage Measurement... 10:46:22 Voltage Offset Correction: v_off_ini = 500.783 uV, v_off_fin = 503.879 uV 10:46:22 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -401.341 uVs 10:46:22 Integrated Voltage Measurement... 10:46:31 Voltage Offset Correction: v_off_ini = 526.693 uV, v_off_fin = 520.116 uV 10:46:31 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 387.623 uVs 10:46:33 I2Y Average and RMS of Measurements VT = -396.855 +- 9.337 uVs I2Y = -36.511 +- 0.859 uTm^2 10:47:01 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:47:06 Integrated Voltage Measurement... 10:47:15 Voltage Offset Correction: v_off_ini = 526.857 uV, v_off_fin = 520.969 uV 10:47:15 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -169.250 uVs 10:47:15 Integrated Voltage Measurement... 10:47:24 Voltage Offset Correction: v_off_ini = 529.580 uV, v_off_fin = 520.101 uV 10:47:24 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.4999 mm dXY = 0.4999 mm VT = 142.828 uVs 10:47:24 Integrated Voltage Measurement... 10:47:33 Voltage Offset Correction: v_off_ini = 528.176 uV, v_off_fin = 515.656 uV 10:47:33 Integrated Voltage Measurement dXYus = -0.4998 mm, dXYds = -0.4999 mm dXY = -0.4999 mm VT = -172.618 uVs 10:47:33 Integrated Voltage Measurement... 10:47:42 Voltage Offset Correction: v_off_ini = 524.265 uV, v_off_fin = 524.368 uV 10:47:42 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 155.178 uVs 10:47:42 Integrated Voltage Measurement... 10:47:51 Voltage Offset Correction: v_off_ini = 524.659 uV, v_off_fin = 521.182 uV 10:47:51 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -173.573 uVs 10:47:51 Integrated Voltage Measurement... 10:48:00 Voltage Offset Correction: v_off_ini = 524.768 uV, v_off_fin = 527.429 uV 10:48:00 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 143.541 uVs 10:48:04 I1X Average and RMS of Measurements VT = -159.498 +- 3.277 uVs I1X = 31.903 +- 0.657 uTm 10:48:06 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:48:11 Integrated Voltage Measurement... 10:48:20 Voltage Offset Correction: v_off_ini = 512.867 uV, v_off_fin = 507.975 uV 10:48:20 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = -193.247 uVs 10:48:20 Integrated Voltage Measurement... 10:48:29 Voltage Offset Correction: v_off_ini = 525.883 uV, v_off_fin = 521.596 uV 10:48:29 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 183.343 uVs 10:48:29 Integrated Voltage Measurement... 10:48:38 Voltage Offset Correction: v_off_ini = 520.045 uV, v_off_fin = 510.745 uV 10:48:38 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -177.864 uVs 10:48:38 Integrated Voltage Measurement... 10:48:47 Voltage Offset Correction: v_off_ini = 518.959 uV, v_off_fin = 536.487 uV 10:48:47 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 186.403 uVs 10:48:47 Integrated Voltage Measurement... 10:48:56 Voltage Offset Correction: v_off_ini = 534.234 uV, v_off_fin = 514.430 uV 10:48:56 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -201.504 uVs 10:48:56 Integrated Voltage Measurement... 10:49:05 Voltage Offset Correction: v_off_ini = 535.687 uV, v_off_fin = 547.243 uV 10:49:05 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 193.351 uVs 10:49:09 I1Y Average and RMS of Measurements VT = -189.285 +- 6.283 uVs I1Y = -37.855 +- 1.257 uTm 10:49:14 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:49:17 Integrated Voltage Measurement... 10:49:26 Voltage Offset Correction: v_off_ini = 534.132 uV, v_off_fin = 527.085 uV 10:49:26 Integrated Voltage Measurement dXYus = -1.0002 mm, dXYds = 0.0000 mm dXY = -1.0002 mm VT = -154.009 uVs 10:49:26 Integrated Voltage Measurement... 10:49:35 Voltage Offset Correction: v_off_ini = 539.619 uV, v_off_fin = 550.312 uV 10:49:35 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 144.727 uVs 10:49:35 Integrated Voltage Measurement... 10:49:44 Voltage Offset Correction: v_off_ini = 537.085 uV, v_off_fin = 530.860 uV 10:49:44 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -132.133 uVs 10:49:44 Integrated Voltage Measurement... 10:49:53 Voltage Offset Correction: v_off_ini = 531.427 uV, v_off_fin = 558.708 uV 10:49:53 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 139.340 uVs 10:49:53 Integrated Voltage Measurement... 10:50:02 Voltage Offset Correction: v_off_ini = 544.566 uV, v_off_fin = 520.669 uV 10:50:02 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -148.083 uVs 10:50:02 Integrated Voltage Measurement... 10:50:11 Voltage Offset Correction: v_off_ini = 549.016 uV, v_off_fin = 560.984 uV 10:50:11 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 153.483 uVs 10:50:13 I2X Average and RMS of Measurements VT = -145.296 +- 6.784 uVs I2X = 13.367 +- 0.624 uTm^2 10:50:18 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:50:21 Integrated Voltage Measurement... 10:50:30 Voltage Offset Correction: v_off_ini = 547.292 uV, v_off_fin = 555.234 uV 10:50:30 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = -0.0001 mm dXY = 0.9999 mm VT = -368.148 uVs 10:50:30 Integrated Voltage Measurement... 10:50:39 Voltage Offset Correction: v_off_ini = 557.174 uV, v_off_fin = 555.029 uV 10:50:39 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0001 mm dXY = -1.0000 mm VT = 373.485 uVs 10:50:39 Integrated Voltage Measurement... 10:50:48 Voltage Offset Correction: v_off_ini = 560.672 uV, v_off_fin = 548.899 uV 10:50:48 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -388.650 uVs 10:50:48 Integrated Voltage Measurement... 10:50:57 Voltage Offset Correction: v_off_ini = 562.747 uV, v_off_fin = 539.717 uV 10:50:57 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 367.953 uVs 10:50:57 Integrated Voltage Measurement... 10:51:06 Voltage Offset Correction: v_off_ini = 552.483 uV, v_off_fin = 558.028 uV 10:51:06 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -389.436 uVs 10:51:06 Integrated Voltage Measurement... 10:51:15 Voltage Offset Correction: v_off_ini = 557.793 uV, v_off_fin = 559.239 uV 10:51:15 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 371.193 uVs 10:51:18 I2Y Average and RMS of Measurements VT = -376.477 +- 4.086 uVs I2Y = -34.636 +- 0.375 uTm^2 10:51:46 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:51:50 Integrated Voltage Measurement... 10:51:59 Voltage Offset Correction: v_off_ini = 560.357 uV, v_off_fin = 556.927 uV 10:51:59 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -178.692 uVs 10:51:59 Integrated Voltage Measurement... 10:52:08 Voltage Offset Correction: v_off_ini = 548.223 uV, v_off_fin = 553.359 uV 10:52:08 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 154.504 uVs 10:52:08 Integrated Voltage Measurement... 10:52:17 Voltage Offset Correction: v_off_ini = 558.105 uV, v_off_fin = 560.953 uV 10:52:17 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -153.605 uVs 10:52:17 Integrated Voltage Measurement... 10:52:26 Voltage Offset Correction: v_off_ini = 551.672 uV, v_off_fin = 572.300 uV 10:52:26 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 123.562 uVs 10:52:26 Integrated Voltage Measurement... 10:52:35 Voltage Offset Correction: v_off_ini = 563.341 uV, v_off_fin = 566.340 uV 10:52:35 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -157.850 uVs 10:52:35 Integrated Voltage Measurement... 10:52:44 Voltage Offset Correction: v_off_ini = 562.600 uV, v_off_fin = 565.166 uV 10:52:44 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 155.533 uVs 10:52:48 I1X Average and RMS of Measurements VT = -153.958 +- 11.599 uVs I1X = 30.792 +- 2.319 uTm 10:52:51 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:52:55 Integrated Voltage Measurement... 10:53:04 Voltage Offset Correction: v_off_ini = 564.527 uV, v_off_fin = 559.302 uV 10:53:04 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = -181.478 uVs 10:53:04 Integrated Voltage Measurement... 10:53:13 Voltage Offset Correction: v_off_ini = 572.664 uV, v_off_fin = 589.506 uV 10:53:13 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 175.761 uVs 10:53:13 Integrated Voltage Measurement... 10:53:22 Voltage Offset Correction: v_off_ini = 583.669 uV, v_off_fin = 572.797 uV 10:53:22 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -194.620 uVs 10:53:22 Integrated Voltage Measurement... 10:53:31 Voltage Offset Correction: v_off_ini = 576.581 uV, v_off_fin = 586.898 uV 10:53:31 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 181.714 uVs 10:53:31 Integrated Voltage Measurement... 10:53:40 Voltage Offset Correction: v_off_ini = 566.262 uV, v_off_fin = 565.376 uV 10:53:40 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -183.369 uVs 10:53:40 Integrated Voltage Measurement... 10:53:49 Voltage Offset Correction: v_off_ini = 576.097 uV, v_off_fin = 593.401 uV 10:53:49 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5002 mm dXY = -0.5001 mm VT = 178.927 uVs 10:53:53 I1Y Average and RMS of Measurements VT = -182.645 +- 4.039 uVs I1Y = -36.528 +- 0.808 uTm 10:53:59 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:54:01 Integrated Voltage Measurement... 10:54:10 Voltage Offset Correction: v_off_ini = 570.516 uV, v_off_fin = 576.328 uV 10:54:10 Integrated Voltage Measurement dXYus = -1.0002 mm, dXYds = 0.0000 mm dXY = -1.0002 mm VT = -160.518 uVs 10:54:10 Integrated Voltage Measurement... 10:54:19 Voltage Offset Correction: v_off_ini = 570.436 uV, v_off_fin = 577.612 uV 10:54:19 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 137.677 uVs 10:54:19 Integrated Voltage Measurement... 10:54:28 Voltage Offset Correction: v_off_ini = 575.679 uV, v_off_fin = 572.936 uV 10:54:28 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -141.047 uVs 10:54:28 Integrated Voltage Measurement... 10:54:37 Voltage Offset Correction: v_off_ini = 579.169 uV, v_off_fin = 592.380 uV 10:54:37 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 141.815 uVs 10:54:37 Integrated Voltage Measurement... 10:54:46 Voltage Offset Correction: v_off_ini = 587.743 uV, v_off_fin = 577.630 uV 10:54:46 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -148.174 uVs 10:54:46 Integrated Voltage Measurement... 10:54:55 Voltage Offset Correction: v_off_ini = 599.372 uV, v_off_fin = 610.820 uV 10:54:55 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 141.156 uVs 10:54:58 I2X Average and RMS of Measurements VT = -145.065 +- 3.143 uVs I2X = 13.345 +- 0.289 uTm^2 10:55:03 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:55:06 Integrated Voltage Measurement... 10:55:15 Voltage Offset Correction: v_off_ini = 592.728 uV, v_off_fin = 588.839 uV 10:55:15 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = -350.876 uVs 10:55:15 Integrated Voltage Measurement... 10:55:24 Voltage Offset Correction: v_off_ini = 592.013 uV, v_off_fin = 602.870 uV 10:55:24 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 367.087 uVs 10:55:24 Integrated Voltage Measurement... 10:55:33 Voltage Offset Correction: v_off_ini = 601.275 uV, v_off_fin = 599.290 uV 10:55:33 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -360.264 uVs 10:55:33 Integrated Voltage Measurement... 10:55:42 Voltage Offset Correction: v_off_ini = 608.585 uV, v_off_fin = 609.161 uV 10:55:42 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 350.212 uVs 10:55:42 Integrated Voltage Measurement... 10:55:51 Voltage Offset Correction: v_off_ini = 602.764 uV, v_off_fin = 596.639 uV 10:55:51 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -347.886 uVs 10:55:51 Integrated Voltage Measurement... 10:56:00 Voltage Offset Correction: v_off_ini = 589.701 uV, v_off_fin = 595.668 uV 10:56:00 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 349.149 uVs 10:56:02 I2Y Average and RMS of Measurements VT = -354.246 +- 4.329 uVs I2Y = -32.591 +- 0.399 uTm^2 10:56:31 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:56:35 Integrated Voltage Measurement... 10:56:44 Voltage Offset Correction: v_off_ini = 608.582 uV, v_off_fin = 596.237 uV 10:56:44 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -147.169 uVs 10:56:44 Integrated Voltage Measurement... 10:56:53 Voltage Offset Correction: v_off_ini = 598.176 uV, v_off_fin = 599.818 uV 10:56:53 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 123.530 uVs 10:56:53 Integrated Voltage Measurement... 10:57:02 Voltage Offset Correction: v_off_ini = 601.802 uV, v_off_fin = 596.319 uV 10:57:02 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -136.772 uVs 10:57:02 Integrated Voltage Measurement... 10:57:11 Voltage Offset Correction: v_off_ini = 596.937 uV, v_off_fin = 596.605 uV 10:57:11 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 133.681 uVs 10:57:11 Integrated Voltage Measurement... 10:57:20 Voltage Offset Correction: v_off_ini = 600.119 uV, v_off_fin = 594.916 uV 10:57:20 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -125.383 uVs 10:57:20 Integrated Voltage Measurement... 10:57:29 Voltage Offset Correction: v_off_ini = 599.175 uV, v_off_fin = 604.580 uV 10:57:29 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 137.743 uVs 10:57:33 I1X Average and RMS of Measurements VT = -134.046 +- 1.757 uVs I1X = 26.810 +- 0.352 uTm 10:57:36 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:57:40 Integrated Voltage Measurement... 10:57:49 Voltage Offset Correction: v_off_ini = 585.167 uV, v_off_fin = 581.703 uV 10:57:49 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -170.834 uVs 10:57:49 Integrated Voltage Measurement... 10:57:58 Voltage Offset Correction: v_off_ini = 581.105 uV, v_off_fin = 606.568 uV 10:57:58 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 192.558 uVs 10:57:58 Integrated Voltage Measurement... 10:58:07 Voltage Offset Correction: v_off_ini = 585.097 uV, v_off_fin = 570.263 uV 10:58:07 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -190.973 uVs 10:58:07 Integrated Voltage Measurement... 10:58:16 Voltage Offset Correction: v_off_ini = 570.335 uV, v_off_fin = 582.725 uV 10:58:16 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 203.443 uVs 10:58:16 Integrated Voltage Measurement... 10:58:25 Voltage Offset Correction: v_off_ini = 583.468 uV, v_off_fin = 568.971 uV 10:58:25 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -212.095 uVs 10:58:25 Integrated Voltage Measurement... 10:58:34 Voltage Offset Correction: v_off_ini = 581.672 uV, v_off_fin = 597.309 uV 10:58:34 Integrated Voltage Measurement dXYus = -0.5001 mm, dXYds = -0.5001 mm dXY = -0.5001 mm VT = 201.817 uVs 10:58:38 I1Y Average and RMS of Measurements VT = -195.287 +- 10.402 uVs I1Y = -39.053 +- 2.079 uTm 10:58:43 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:58:46 Integrated Voltage Measurement... 10:58:55 Voltage Offset Correction: v_off_ini = 576.126 uV, v_off_fin = 569.963 uV 10:58:55 Integrated Voltage Measurement dXYus = -1.0002 mm, dXYds = 0.0000 mm dXY = -1.0002 mm VT = -136.444 uVs 10:58:55 Integrated Voltage Measurement... 10:59:04 Voltage Offset Correction: v_off_ini = 583.959 uV, v_off_fin = 588.399 uV 10:59:04 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 121.268 uVs 10:59:04 Integrated Voltage Measurement... 10:59:13 Voltage Offset Correction: v_off_ini = 584.391 uV, v_off_fin = 575.241 uV 10:59:13 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -137.479 uVs 10:59:13 Integrated Voltage Measurement... 10:59:22 Voltage Offset Correction: v_off_ini = 573.194 uV, v_off_fin = 569.948 uV 10:59:22 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 121.723 uVs 10:59:22 Integrated Voltage Measurement... 10:59:31 Voltage Offset Correction: v_off_ini = 566.877 uV, v_off_fin = 556.928 uV 10:59:31 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -139.203 uVs 10:59:31 Integrated Voltage Measurement... 10:59:40 Voltage Offset Correction: v_off_ini = 561.500 uV, v_off_fin = 578.950 uV 10:59:40 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 121.647 uVs 10:59:43 I2X Average and RMS of Measurements VT = -129.627 +- 0.641 uVs I2X = 11.926 +- 0.059 uTm^2 10:59:48 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:59:50 Integrated Voltage Measurement... 10:59:59 Voltage Offset Correction: v_off_ini = 587.263 uV, v_off_fin = 575.953 uV 10:59:59 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -343.618 uVs 10:59:59 Integrated Voltage Measurement... 11:00:08 Voltage Offset Correction: v_off_ini = 567.379 uV, v_off_fin = 575.038 uV 11:00:08 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 348.810 uVs 11:00:08 Integrated Voltage Measurement... 11:00:17 Voltage Offset Correction: v_off_ini = 586.220 uV, v_off_fin = 580.704 uV 11:00:17 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -350.946 uVs 11:00:17 Integrated Voltage Measurement... 11:00:26 Voltage Offset Correction: v_off_ini = 566.843 uV, v_off_fin = 572.195 uV 11:00:26 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 340.761 uVs 11:00:27 Integrated Voltage Measurement... 11:00:36 Voltage Offset Correction: v_off_ini = 574.918 uV, v_off_fin = 579.576 uV 11:00:36 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -342.878 uVs 11:00:36 Integrated Voltage Measurement... 11:00:45 Voltage Offset Correction: v_off_ini = 585.053 uV, v_off_fin = 576.063 uV 11:00:45 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 340.895 uVs 11:00:47 I2Y Average and RMS of Measurements VT = -344.651 +- 1.960 uVs I2Y = -31.708 +- 0.180 uTm^2 11:01:15 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:01:20 Integrated Voltage Measurement... 11:01:29 Voltage Offset Correction: v_off_ini = 605.083 uV, v_off_fin = 582.751 uV 11:01:29 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -150.875 uVs 11:01:29 Integrated Voltage Measurement... 11:01:38 Voltage Offset Correction: v_off_ini = 604.046 uV, v_off_fin = 602.673 uV 11:01:38 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 103.799 uVs 11:01:38 Integrated Voltage Measurement... 11:01:47 Voltage Offset Correction: v_off_ini = 588.486 uV, v_off_fin = 593.708 uV 11:01:47 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -130.513 uVs 11:01:47 Integrated Voltage Measurement... 11:01:56 Voltage Offset Correction: v_off_ini = 598.214 uV, v_off_fin = 601.883 uV 11:01:56 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 118.652 uVs 11:01:56 Integrated Voltage Measurement... 11:02:05 Voltage Offset Correction: v_off_ini = 602.666 uV, v_off_fin = 597.028 uV 11:02:05 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -128.713 uVs 11:02:05 Integrated Voltage Measurement... 11:02:14 Voltage Offset Correction: v_off_ini = 607.675 uV, v_off_fin = 607.583 uV 11:02:14 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 132.245 uVs 11:02:18 I1X Average and RMS of Measurements VT = -127.466 +- 2.409 uVs I1X = 25.494 +- 0.481 uTm 11:02:21 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:02:25 Integrated Voltage Measurement... 11:02:34 Voltage Offset Correction: v_off_ini = 600.498 uV, v_off_fin = 590.773 uV 11:02:34 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -176.358 uVs 11:02:34 Integrated Voltage Measurement... 11:02:43 Voltage Offset Correction: v_off_ini = 588.196 uV, v_off_fin = 614.769 uV 11:02:43 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 197.396 uVs 11:02:43 Integrated Voltage Measurement... 11:02:52 Voltage Offset Correction: v_off_ini = 606.759 uV, v_off_fin = 591.433 uV 11:02:52 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -194.603 uVs 11:02:52 Integrated Voltage Measurement... 11:03:01 Voltage Offset Correction: v_off_ini = 605.869 uV, v_off_fin = 622.716 uV 11:03:01 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 195.462 uVs 11:03:01 Integrated Voltage Measurement... 11:03:10 Voltage Offset Correction: v_off_ini = 610.581 uV, v_off_fin = 599.468 uV 11:03:10 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -184.075 uVs 11:03:10 Integrated Voltage Measurement... 11:03:19 Voltage Offset Correction: v_off_ini = 619.385 uV, v_off_fin = 622.831 uV 11:03:19 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 191.943 uVs 11:03:23 I1Y Average and RMS of Measurements VT = -189.973 +- 3.607 uVs I1Y = -37.991 +- 0.721 uTm 11:03:28 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:03:31 Integrated Voltage Measurement... 11:03:40 Voltage Offset Correction: v_off_ini = 622.191 uV, v_off_fin = 601.935 uV 11:03:40 Integrated Voltage Measurement dXYus = -1.0002 mm, dXYds = 0.0000 mm dXY = -1.0002 mm VT = -120.772 uVs 11:03:40 Integrated Voltage Measurement... 11:03:49 Voltage Offset Correction: v_off_ini = 610.035 uV, v_off_fin = 623.106 uV 11:03:49 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 111.581 uVs 11:03:49 Integrated Voltage Measurement... 11:03:58 Voltage Offset Correction: v_off_ini = 620.956 uV, v_off_fin = 609.506 uV 11:03:58 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -113.445 uVs 11:03:58 Integrated Voltage Measurement... 11:04:07 Voltage Offset Correction: v_off_ini = 612.425 uV, v_off_fin = 627.092 uV 11:04:07 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 115.543 uVs 11:04:07 Integrated Voltage Measurement... 11:04:16 Voltage Offset Correction: v_off_ini = 622.402 uV, v_off_fin = 618.020 uV 11:04:16 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -135.603 uVs 11:04:16 Integrated Voltage Measurement... 11:04:25 Voltage Offset Correction: v_off_ini = 620.414 uV, v_off_fin = 640.423 uV 11:04:25 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 110.648 uVs 11:04:27 I2X Average and RMS of Measurements VT = -117.932 +- 3.736 uVs I2X = 10.850 +- 0.344 uTm^2 11:04:33 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:04:35 Integrated Voltage Measurement... 11:04:44 Voltage Offset Correction: v_off_ini = 622.260 uV, v_off_fin = 621.122 uV 11:04:44 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -319.186 uVs 11:04:44 Integrated Voltage Measurement... 11:04:53 Voltage Offset Correction: v_off_ini = 613.299 uV, v_off_fin = 602.362 uV 11:04:53 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 331.698 uVs 11:04:53 Integrated Voltage Measurement... 11:05:02 Voltage Offset Correction: v_off_ini = 614.289 uV, v_off_fin = 608.933 uV 11:05:02 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -308.805 uVs 11:05:02 Integrated Voltage Measurement... 11:05:11 Voltage Offset Correction: v_off_ini = 599.429 uV, v_off_fin = 614.350 uV 11:05:11 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 319.393 uVs 11:05:11 Integrated Voltage Measurement... 11:05:20 Voltage Offset Correction: v_off_ini = 621.684 uV, v_off_fin = 609.808 uV 11:05:20 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -323.764 uVs 11:05:20 Integrated Voltage Measurement... 11:05:29 Voltage Offset Correction: v_off_ini = 615.528 uV, v_off_fin = 620.615 uV 11:05:29 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 309.034 uVs 11:05:32 I2Y Average and RMS of Measurements VT = -318.647 +- 4.896 uVs I2Y = -29.315 +- 0.450 uTm^2