SLAC Magnetic Measurements Date: 07-17-2025 Time: 15:15:03 Project: LCLS-II-HE Magnet Type: Phase Shifter Magnet Name: HE_SXPS_16353 Dataset Number: 1 Run Purpose: Tuning Run Type: Stretched Wire Run Number: 7 Measurement Device: Operator: YL Comment: 15:15:11 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 15:15:13 Motion Parameters, X Controller, Axis 1 acc = 0.001000 m/s/s dcl = 0.001000 m/s/s vel = 0.001000 m/s pos = 0.000000 m kp = 188.600000 ki = 820.000000 kd = 491.400000 db = 0.000000 15:15:15 Motion Parameters, X Controller, Axis 2 acc = 0.001000 m/s/s dcl = 0.001000 m/s/s vel = 0.001000 m/s pos = 0.000000 m kp = 188.600000 ki = 820.000000 kd = 491.400000 db = 0.000000 15:15:16 Motion Parameters, Y Controller, Axis 1 acc = 0.001000 m/s/s dcl = 0.001000 m/s/s vel = 0.001000 m/s pos = 0.000000 m kp = 188.600000 ki = 820.000000 kd = 491.400000 db = 0.000000 15:15:18 Motion Parameters, Y Controller, Axis 2 acc = 0.001000 m/s/s dcl = 0.001000 m/s/s vel = 0.001000 m/s pos = 0.000000 m kp = 188.600000 ki = 820.000000 kd = 491.400000 db = 0.000000 15:15:28 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 15:15:33 Integrated Voltage Measurement... 15:15:42 Voltage Offset Correction: v_off_ini = 147.201 uV, v_off_fin = 142.818 uV 15:15:42 Integrated Voltage Measurement dXYus = -0.4998 mm, dXYds = -0.4998 mm dXY = -0.4998 mm VT = -50.205 uVs 15:15:42 Integrated Voltage Measurement... 15:15:51 Voltage Offset Correction: v_off_ini = 137.614 uV, v_off_fin = 144.083 uV 15:15:51 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 19.369 uVs 15:15:51 Integrated Voltage Measurement... 15:16:00 Voltage Offset Correction: v_off_ini = 136.803 uV, v_off_fin = 133.549 uV 15:16:00 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -21.980 uVs 15:16:00 Integrated Voltage Measurement... 15:16:09 Voltage Offset Correction: v_off_ini = 157.215 uV, v_off_fin = 168.923 uV 15:16:09 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 36.519 uVs 15:16:09 Integrated Voltage Measurement... 15:16:18 Voltage Offset Correction: v_off_ini = 166.218 uV, v_off_fin = 151.759 uV 15:16:18 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -27.078 uVs 15:16:18 Integrated Voltage Measurement... 15:16:26 Voltage Offset Correction: v_off_ini = 156.549 uV, v_off_fin = 171.741 uV 15:16:27 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 29.491 uVs 15:16:31 I1X Average and RMS of Measurements VT = -30.774 +- 2.865 uVs I1X = 6.156 +- 0.574 uTm 15:16:33 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 15:16:37 Integrated Voltage Measurement... 15:16:46 Voltage Offset Correction: v_off_ini = 183.897 uV, v_off_fin = 177.791 uV 15:16:46 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -158.569 uVs 15:16:46 Integrated Voltage Measurement... 15:16:55 Voltage Offset Correction: v_off_ini = 179.901 uV, v_off_fin = 205.643 uV 15:16:55 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 183.652 uVs 15:16:55 Integrated Voltage Measurement... 15:17:04 Voltage Offset Correction: v_off_ini = 189.919 uV, v_off_fin = 179.012 uV 15:17:04 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -155.746 uVs 15:17:04 Integrated Voltage Measurement... 15:17:13 Voltage Offset Correction: v_off_ini = 196.999 uV, v_off_fin = 209.691 uV 15:17:13 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 178.323 uVs 15:17:13 Integrated Voltage Measurement... 15:17:22 Voltage Offset Correction: v_off_ini = 198.280 uV, v_off_fin = 184.356 uV 15:17:22 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -168.521 uVs 15:17:22 Integrated Voltage Measurement... 15:17:31 Voltage Offset Correction: v_off_ini = 192.948 uV, v_off_fin = 205.389 uV 15:17:31 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 183.931 uVs 15:17:36 I1Y Average and RMS of Measurements VT = -171.457 +- 3.761 uVs I1Y = -34.288 +- 0.752 uTm 15:17:41 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 15:17:43 Integrated Voltage Measurement... 15:17:52 Voltage Offset Correction: v_off_ini = 193.964 uV, v_off_fin = 192.376 uV 15:17:52 Integrated Voltage Measurement dXYus = -1.0002 mm, dXYds = 0.0000 mm dXY = -1.0002 mm VT = 2.116 uVs 15:17:52 Integrated Voltage Measurement... 15:18:01 Voltage Offset Correction: v_off_ini = 199.090 uV, v_off_fin = 206.634 uV 15:18:01 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -12.693 uVs 15:18:01 Integrated Voltage Measurement... 15:18:10 Voltage Offset Correction: v_off_ini = 196.941 uV, v_off_fin = 198.517 uV 15:18:10 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 12.495 uVs 15:18:10 Integrated Voltage Measurement... 15:18:19 Voltage Offset Correction: v_off_ini = 217.587 uV, v_off_fin = 230.740 uV 15:18:19 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -8.365 uVs 15:18:19 Integrated Voltage Measurement... 15:18:28 Voltage Offset Correction: v_off_ini = 223.317 uV, v_off_fin = 214.320 uV 15:18:28 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 20.984 uVs 15:18:28 Integrated Voltage Measurement... 15:18:37 Voltage Offset Correction: v_off_ini = 221.621 uV, v_off_fin = 230.964 uV 15:18:37 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -5.517 uVs 15:18:40 I2X Average and RMS of Measurements VT = 10.362 +- 2.387 uVs I2X = -0.953 +- 0.220 uTm^2 15:18:45 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 15:18:47 Integrated Voltage Measurement... 15:18:56 Voltage Offset Correction: v_off_ini = 214.370 uV, v_off_fin = 218.619 uV 15:18:56 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -495.686 uVs 15:18:56 Integrated Voltage Measurement... 15:19:05 Voltage Offset Correction: v_off_ini = 227.846 uV, v_off_fin = 222.054 uV 15:19:05 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = 520.790 uVs 15:19:05 Integrated Voltage Measurement... 15:19:14 Voltage Offset Correction: v_off_ini = 228.782 uV, v_off_fin = 242.673 uV 15:19:14 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -518.231 uVs 15:19:14 Integrated Voltage Measurement... 15:19:23 Voltage Offset Correction: v_off_ini = 244.136 uV, v_off_fin = 243.085 uV 15:19:23 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 494.849 uVs 15:19:23 Integrated Voltage Measurement... 15:19:32 Voltage Offset Correction: v_off_ini = 255.813 uV, v_off_fin = 253.378 uV 15:19:32 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -506.152 uVs 15:19:32 Integrated Voltage Measurement... 15:19:41 Voltage Offset Correction: v_off_ini = 249.438 uV, v_off_fin = 244.964 uV 15:19:41 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 503.473 uVs 15:19:44 I2Y Average and RMS of Measurements VT = -506.530 +- 1.398 uVs I2Y = -46.602 +- 0.130 uTm^2 15:19:55 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 15:19:59 Integrated Voltage Measurement... 15:20:08 Voltage Offset Correction: v_off_ini = 248.815 uV, v_off_fin = 252.400 uV 15:20:08 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -29.918 uVs 15:20:08 Integrated Voltage Measurement... 15:20:17 Voltage Offset Correction: v_off_ini = 243.918 uV, v_off_fin = 258.605 uV 15:20:17 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 53.060 uVs 15:20:17 Integrated Voltage Measurement... 15:20:26 Voltage Offset Correction: v_off_ini = 253.534 uV, v_off_fin = 257.518 uV 15:20:26 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -61.080 uVs 15:20:26 Integrated Voltage Measurement... 15:20:35 Voltage Offset Correction: v_off_ini = 247.253 uV, v_off_fin = 262.390 uV 15:20:35 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 52.591 uVs 15:20:35 Integrated Voltage Measurement... 15:20:44 Voltage Offset Correction: v_off_ini = 262.863 uV, v_off_fin = 255.528 uV 15:20:44 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -23.637 uVs 15:20:44 Integrated Voltage Measurement... 15:20:53 Voltage Offset Correction: v_off_ini = 271.921 uV, v_off_fin = 259.801 uV 15:20:53 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 36.274 uVs 15:20:57 I1X Average and RMS of Measurements VT = -42.760 +- 11.010 uVs I1X = 8.552 +- 2.202 uTm 15:21:00 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 15:21:04 Integrated Voltage Measurement... 15:21:13 Voltage Offset Correction: v_off_ini = 255.834 uV, v_off_fin = 248.793 uV 15:21:13 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -158.050 uVs 15:21:13 Integrated Voltage Measurement... 15:21:22 Voltage Offset Correction: v_off_ini = 255.157 uV, v_off_fin = 261.457 uV 15:21:22 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 186.726 uVs 15:21:22 Integrated Voltage Measurement... 15:21:31 Voltage Offset Correction: v_off_ini = 245.424 uV, v_off_fin = 242.800 uV 15:21:31 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -169.107 uVs 15:21:31 Integrated Voltage Measurement... 15:21:40 Voltage Offset Correction: v_off_ini = 264.443 uV, v_off_fin = 276.161 uV 15:21:40 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 182.469 uVs 15:21:40 Integrated Voltage Measurement... 15:21:49 Voltage Offset Correction: v_off_ini = 257.317 uV, v_off_fin = 264.634 uV 15:21:49 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -172.024 uVs 15:21:49 Integrated Voltage Measurement... 15:21:58 Voltage Offset Correction: v_off_ini = 258.827 uV, v_off_fin = 259.730 uV 15:21:58 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 208.669 uVs 15:22:02 I1Y Average and RMS of Measurements VT = -179.507 +- 7.789 uVs I1Y = -35.898 +- 1.557 uTm 15:22:07 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 15:22:10 Integrated Voltage Measurement... 15:22:19 Voltage Offset Correction: v_off_ini = 254.982 uV, v_off_fin = 244.267 uV 15:22:19 Integrated Voltage Measurement dXYus = -1.0002 mm, dXYds = 0.0000 mm dXY = -1.0002 mm VT = 5.619 uVs 15:22:19 Integrated Voltage Measurement... 15:22:28 Voltage Offset Correction: v_off_ini = 242.584 uV, v_off_fin = 260.726 uV 15:22:28 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 14.981 uVs 15:22:28 Integrated Voltage Measurement... 15:22:37 Voltage Offset Correction: v_off_ini = 233.988 uV, v_off_fin = 241.680 uV 15:22:37 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 28.566 uVs 15:22:37 Integrated Voltage Measurement... 15:22:46 Voltage Offset Correction: v_off_ini = 248.910 uV, v_off_fin = 252.330 uV 15:22:46 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 5.296 uVs 15:22:46 Integrated Voltage Measurement... 15:22:55 Voltage Offset Correction: v_off_ini = 247.384 uV, v_off_fin = 242.791 uV 15:22:55 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -5.469 uVs 15:22:55 Integrated Voltage Measurement... 15:23:04 Voltage Offset Correction: v_off_ini = 252.108 uV, v_off_fin = 266.443 uV 15:23:04 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 15.781 uVs 15:23:07 I2X Average and RMS of Measurements VT = -1.224 +- 9.411 uVs I2X = 0.113 +- 0.866 uTm^2 15:23:12 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 15:23:14 Integrated Voltage Measurement... 15:23:23 Voltage Offset Correction: v_off_ini = 247.796 uV, v_off_fin = 252.358 uV 15:23:23 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -501.703 uVs 15:23:23 Integrated Voltage Measurement... 15:23:32 Voltage Offset Correction: v_off_ini = 254.703 uV, v_off_fin = 253.244 uV 15:23:32 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 495.602 uVs 15:23:32 Integrated Voltage Measurement... 15:23:41 Voltage Offset Correction: v_off_ini = 253.662 uV, v_off_fin = 240.802 uV 15:23:41 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = -519.407 uVs 15:23:41 Integrated Voltage Measurement... 15:23:50 Voltage Offset Correction: v_off_ini = 249.550 uV, v_off_fin = 257.512 uV 15:23:50 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 480.257 uVs 15:23:50 Integrated Voltage Measurement... 15:23:59 Voltage Offset Correction: v_off_ini = 238.339 uV, v_off_fin = 239.998 uV 15:23:59 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -525.592 uVs 15:23:59 Integrated Voltage Measurement... 15:24:08 Voltage Offset Correction: v_off_ini = 245.724 uV, v_off_fin = 239.080 uV 15:24:08 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 496.287 uVs 15:24:11 I2Y Average and RMS of Measurements VT = -503.141 +- 5.535 uVs I2Y = -46.290 +- 0.509 uTm^2 15:24:22 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 15:24:26 Integrated Voltage Measurement... 15:24:35 Voltage Offset Correction: v_off_ini = 239.707 uV, v_off_fin = 257.737 uV 15:24:35 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -28.674 uVs 15:24:35 Integrated Voltage Measurement... 15:24:44 Voltage Offset Correction: v_off_ini = 257.800 uV, v_off_fin = 262.299 uV 15:24:44 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 44.626 uVs 15:24:44 Integrated Voltage Measurement... 15:24:53 Voltage Offset Correction: v_off_ini = 267.163 uV, v_off_fin = 261.708 uV 15:24:53 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -50.305 uVs 15:24:53 Integrated Voltage Measurement... 15:25:02 Voltage Offset Correction: v_off_ini = 266.563 uV, v_off_fin = 277.089 uV 15:25:02 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 45.557 uVs 15:25:02 Integrated Voltage Measurement... 15:25:11 Voltage Offset Correction: v_off_ini = 259.900 uV, v_off_fin = 274.184 uV 15:25:11 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -63.956 uVs 15:25:11 Integrated Voltage Measurement... 15:25:20 Voltage Offset Correction: v_off_ini = 270.107 uV, v_off_fin = 284.069 uV 15:25:20 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 58.073 uVs 15:25:24 I1X Average and RMS of Measurements VT = -48.532 +- 9.956 uVs I1X = 9.706 +- 1.991 uTm 15:25:27 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 15:25:31 Integrated Voltage Measurement... 15:25:40 Voltage Offset Correction: v_off_ini = 263.743 uV, v_off_fin = 269.027 uV 15:25:40 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -168.778 uVs 15:25:40 Integrated Voltage Measurement... 15:25:49 Voltage Offset Correction: v_off_ini = 266.336 uV, v_off_fin = 300.347 uV 15:25:49 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5002 mm dXY = -0.5001 mm VT = 184.159 uVs 15:25:49 Integrated Voltage Measurement... 15:25:58 Voltage Offset Correction: v_off_ini = 281.432 uV, v_off_fin = 271.628 uV 15:25:58 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5002 mm dXY = 0.5001 mm VT = -167.685 uVs 15:25:58 Integrated Voltage Measurement... 15:26:07 Voltage Offset Correction: v_off_ini = 262.796 uV, v_off_fin = 283.488 uV 15:26:07 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 202.683 uVs 15:26:07 Integrated Voltage Measurement... 15:26:16 Voltage Offset Correction: v_off_ini = 286.098 uV, v_off_fin = 273.242 uV 15:26:16 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -175.664 uVs 15:26:16 Integrated Voltage Measurement... 15:26:25 Voltage Offset Correction: v_off_ini = 280.160 uV, v_off_fin = 297.455 uV 15:26:25 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 202.042 uVs 15:26:29 I1Y Average and RMS of Measurements VT = -183.502 +- 5.194 uVs I1Y = -36.696 +- 1.039 uTm 15:26:34 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 15:26:37 Integrated Voltage Measurement... 15:26:46 Voltage Offset Correction: v_off_ini = 304.653 uV, v_off_fin = 281.584 uV 15:26:46 Integrated Voltage Measurement dXYus = -1.0002 mm, dXYds = 0.0000 mm dXY = -1.0002 mm VT = -8.472 uVs 15:26:46 Integrated Voltage Measurement... 15:26:55 Voltage Offset Correction: v_off_ini = 295.737 uV, v_off_fin = 300.139 uV 15:26:55 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 16.632 uVs 15:26:55 Integrated Voltage Measurement... 15:27:04 Voltage Offset Correction: v_off_ini = 281.894 uV, v_off_fin = 286.069 uV 15:27:04 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -30.588 uVs 15:27:04 Integrated Voltage Measurement... 15:27:13 Voltage Offset Correction: v_off_ini = 281.940 uV, v_off_fin = 307.118 uV 15:27:13 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 29.250 uVs 15:27:13 Integrated Voltage Measurement... 15:27:22 Voltage Offset Correction: v_off_ini = 293.113 uV, v_off_fin = 296.095 uV 15:27:22 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -25.183 uVs 15:27:22 Integrated Voltage Measurement... 15:27:31 Voltage Offset Correction: v_off_ini = 298.215 uV, v_off_fin = 300.383 uV 15:27:31 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 22.677 uVs 15:27:33 I2X Average and RMS of Measurements VT = -22.134 +- 7.203 uVs I2X = 2.036 +- 0.663 uTm^2 15:27:39 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 15:27:41 Integrated Voltage Measurement... 15:27:50 Voltage Offset Correction: v_off_ini = 306.296 uV, v_off_fin = 293.191 uV 15:27:50 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -505.607 uVs 15:27:50 Integrated Voltage Measurement... 15:27:59 Voltage Offset Correction: v_off_ini = 288.019 uV, v_off_fin = 293.709 uV 15:27:59 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 492.912 uVs 15:27:59 Integrated Voltage Measurement... 15:28:08 Voltage Offset Correction: v_off_ini = 293.357 uV, v_off_fin = 279.148 uV 15:28:08 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -490.627 uVs 15:28:08 Integrated Voltage Measurement... 15:28:17 Voltage Offset Correction: v_off_ini = 283.108 uV, v_off_fin = 300.080 uV 15:28:17 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 483.785 uVs 15:28:17 Integrated Voltage Measurement... 15:28:26 Voltage Offset Correction: v_off_ini = 285.406 uV, v_off_fin = 291.368 uV 15:28:26 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -503.109 uVs 15:28:26 Integrated Voltage Measurement... 15:28:35 Voltage Offset Correction: v_off_ini = 278.374 uV, v_off_fin = 289.131 uV 15:28:35 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 514.046 uVs 15:28:38 I2Y Average and RMS of Measurements VT = -498.348 +- 8.749 uVs I2Y = -45.848 +- 0.805 uTm^2 15:28:48 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 15:28:53 Integrated Voltage Measurement... 15:29:02 Voltage Offset Correction: v_off_ini = 315.490 uV, v_off_fin = 312.525 uV 15:29:02 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -59.595 uVs 15:29:02 Integrated Voltage Measurement... 15:29:11 Voltage Offset Correction: v_off_ini = 314.447 uV, v_off_fin = 308.236 uV 15:29:11 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 76.004 uVs 15:29:11 Integrated Voltage Measurement... 15:29:20 Voltage Offset Correction: v_off_ini = 307.373 uV, v_off_fin = 310.686 uV 15:29:20 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -61.230 uVs 15:29:20 Integrated Voltage Measurement... 15:29:29 Voltage Offset Correction: v_off_ini = 304.614 uV, v_off_fin = 309.862 uV 15:29:29 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 72.813 uVs 15:29:29 Integrated Voltage Measurement... 15:29:38 Voltage Offset Correction: v_off_ini = 306.414 uV, v_off_fin = 304.629 uV 15:29:38 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -65.148 uVs 15:29:38 Integrated Voltage Measurement... 15:29:47 Voltage Offset Correction: v_off_ini = 312.092 uV, v_off_fin = 305.307 uV 15:29:47 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 71.047 uVs 15:29:51 I1X Average and RMS of Measurements VT = -67.639 +- 0.454 uVs I1X = 13.528 +- 0.091 uTm 15:29:54 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 15:29:58 Integrated Voltage Measurement... 15:30:07 Voltage Offset Correction: v_off_ini = 312.606 uV, v_off_fin = 298.884 uV 15:30:07 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -167.258 uVs 15:30:07 Integrated Voltage Measurement... 15:30:16 Voltage Offset Correction: v_off_ini = 292.491 uV, v_off_fin = 313.778 uV 15:30:16 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 202.966 uVs 15:30:16 Integrated Voltage Measurement... 15:30:25 Voltage Offset Correction: v_off_ini = 301.717 uV, v_off_fin = 297.776 uV 15:30:25 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -168.994 uVs 15:30:25 Integrated Voltage Measurement... 15:30:34 Voltage Offset Correction: v_off_ini = 304.516 uV, v_off_fin = 314.692 uV 15:30:34 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 169.229 uVs 15:30:34 Integrated Voltage Measurement... 15:30:43 Voltage Offset Correction: v_off_ini = 304.225 uV, v_off_fin = 282.831 uV 15:30:43 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -159.088 uVs 15:30:43 Integrated Voltage Measurement... 15:30:52 Voltage Offset Correction: v_off_ini = 292.918 uV, v_off_fin = 321.425 uV 15:30:52 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 176.654 uVs 15:30:56 I1Y Average and RMS of Measurements VT = -174.032 +- 7.851 uVs I1Y = -34.803 +- 1.571 uTm 15:31:01 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 15:31:04 Integrated Voltage Measurement... 15:31:13 Voltage Offset Correction: v_off_ini = 303.851 uV, v_off_fin = 282.387 uV 15:31:13 Integrated Voltage Measurement dXYus = -1.0002 mm, dXYds = 0.0000 mm dXY = -1.0002 mm VT = -29.054 uVs 15:31:13 Integrated Voltage Measurement... 15:31:22 Voltage Offset Correction: v_off_ini = 299.920 uV, v_off_fin = 315.858 uV 15:31:22 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 43.407 uVs 15:31:22 Integrated Voltage Measurement... 15:31:31 Voltage Offset Correction: v_off_ini = 307.203 uV, v_off_fin = 292.485 uV 15:31:31 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -52.938 uVs 15:31:31 Integrated Voltage Measurement... 15:31:40 Voltage Offset Correction: v_off_ini = 308.484 uV, v_off_fin = 314.075 uV 15:31:40 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 19.856 uVs 15:31:40 Integrated Voltage Measurement... 15:31:49 Voltage Offset Correction: v_off_ini = 313.310 uV, v_off_fin = 302.395 uV 15:31:49 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -30.437 uVs 15:31:49 Integrated Voltage Measurement... 15:31:58 Voltage Offset Correction: v_off_ini = 319.076 uV, v_off_fin = 315.866 uV 15:31:58 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 1.458 uVs 15:32:00 I2X Average and RMS of Measurements VT = -29.525 +- 9.601 uVs I2X = 2.716 +- 0.883 uTm^2 15:32:05 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 15:32:08 Integrated Voltage Measurement... 15:32:17 Voltage Offset Correction: v_off_ini = 301.993 uV, v_off_fin = 300.207 uV 15:32:17 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -485.891 uVs 15:32:17 Integrated Voltage Measurement... 15:32:26 Voltage Offset Correction: v_off_ini = 294.736 uV, v_off_fin = 297.439 uV 15:32:26 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 507.452 uVs 15:32:26 Integrated Voltage Measurement... 15:32:35 Voltage Offset Correction: v_off_ini = 300.578 uV, v_off_fin = 302.476 uV 15:32:35 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -492.467 uVs 15:32:35 Integrated Voltage Measurement... 15:32:44 Voltage Offset Correction: v_off_ini = 307.068 uV, v_off_fin = 299.128 uV 15:32:44 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 506.360 uVs 15:32:44 Integrated Voltage Measurement... 15:32:53 Voltage Offset Correction: v_off_ini = 294.069 uV, v_off_fin = 295.691 uV 15:32:53 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -493.921 uVs 15:32:53 Integrated Voltage Measurement... 15:33:02 Voltage Offset Correction: v_off_ini = 320.515 uV, v_off_fin = 312.824 uV 15:33:02 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 504.223 uVs 15:33:05 I2Y Average and RMS of Measurements VT = -498.386 +- 1.220 uVs I2Y = -45.851 +- 0.112 uTm^2 15:33:15 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 15:33:20 Integrated Voltage Measurement... 15:33:28 Voltage Offset Correction: v_off_ini = 322.766 uV, v_off_fin = 317.020 uV 15:33:29 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -85.382 uVs 15:33:29 Integrated Voltage Measurement... 15:33:38 Voltage Offset Correction: v_off_ini = 310.286 uV, v_off_fin = 321.259 uV 15:33:38 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 65.150 uVs 15:33:38 Integrated Voltage Measurement... 15:33:47 Voltage Offset Correction: v_off_ini = 330.839 uV, v_off_fin = 323.260 uV 15:33:47 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -65.684 uVs 15:33:47 Integrated Voltage Measurement... 15:33:56 Voltage Offset Correction: v_off_ini = 321.773 uV, v_off_fin = 332.825 uV 15:33:56 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 56.686 uVs 15:33:56 Integrated Voltage Measurement... 15:34:05 Voltage Offset Correction: v_off_ini = 349.804 uV, v_off_fin = 335.920 uV 15:34:05 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -83.221 uVs 15:34:05 Integrated Voltage Measurement... 15:34:14 Voltage Offset Correction: v_off_ini = 333.100 uV, v_off_fin = 334.843 uV 15:34:14 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 64.016 uVs 15:34:18 I1X Average and RMS of Measurements VT = -70.023 +- 6.286 uVs I1X = 14.005 +- 1.257 uTm 15:34:20 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 15:34:25 Integrated Voltage Measurement... 15:34:34 Voltage Offset Correction: v_off_ini = 320.974 uV, v_off_fin = 335.673 uV 15:34:34 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -163.913 uVs 15:34:34 Integrated Voltage Measurement... 15:34:43 Voltage Offset Correction: v_off_ini = 333.961 uV, v_off_fin = 355.984 uV 15:34:43 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 184.995 uVs 15:34:43 Integrated Voltage Measurement... 15:34:52 Voltage Offset Correction: v_off_ini = 331.826 uV, v_off_fin = 319.807 uV 15:34:52 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -165.664 uVs 15:34:52 Integrated Voltage Measurement... 15:35:01 Voltage Offset Correction: v_off_ini = 338.152 uV, v_off_fin = 354.756 uV 15:35:01 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 192.351 uVs 15:35:01 Integrated Voltage Measurement... 15:35:10 Voltage Offset Correction: v_off_ini = 334.872 uV, v_off_fin = 331.599 uV 15:35:10 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -174.187 uVs 15:35:10 Integrated Voltage Measurement... 15:35:19 Voltage Offset Correction: v_off_ini = 343.224 uV, v_off_fin = 354.754 uV 15:35:19 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 211.437 uVs 15:35:23 I1Y Average and RMS of Measurements VT = -182.091 +- 7.805 uVs I1Y = -36.416 +- 1.560 uTm 15:35:28 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 15:35:31 Integrated Voltage Measurement... 15:35:40 Voltage Offset Correction: v_off_ini = 353.571 uV, v_off_fin = 341.553 uV 15:35:40 Integrated Voltage Measurement dXYus = -1.0002 mm, dXYds = 0.0000 mm dXY = -1.0002 mm VT = -44.069 uVs 15:35:40 Integrated Voltage Measurement... 15:35:49 Voltage Offset Correction: v_off_ini = 355.085 uV, v_off_fin = 368.788 uV 15:35:49 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 15.743 uVs 15:35:49 Integrated Voltage Measurement... 15:35:58 Voltage Offset Correction: v_off_ini = 352.369 uV, v_off_fin = 339.186 uV 15:35:58 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -48.247 uVs 15:35:58 Integrated Voltage Measurement... 15:36:07 Voltage Offset Correction: v_off_ini = 349.231 uV, v_off_fin = 356.765 uV 15:36:07 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 42.200 uVs 15:36:07 Integrated Voltage Measurement... 15:36:16 Voltage Offset Correction: v_off_ini = 346.284 uV, v_off_fin = 339.973 uV 15:36:16 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -30.568 uVs 15:36:16 Integrated Voltage Measurement... 15:36:25 Voltage Offset Correction: v_off_ini = 349.434 uV, v_off_fin = 371.395 uV 15:36:25 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 26.699 uVs 15:36:27 I2X Average and RMS of Measurements VT = -34.588 +- 7.539 uVs I2X = 3.182 +- 0.694 uTm^2 15:36:32 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 15:36:35 Integrated Voltage Measurement... 15:36:44 Voltage Offset Correction: v_off_ini = 363.889 uV, v_off_fin = 370.515 uV 15:36:44 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -495.201 uVs 15:36:44 Integrated Voltage Measurement... 15:36:53 Voltage Offset Correction: v_off_ini = 361.446 uV, v_off_fin = 366.936 uV 15:36:53 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = 503.315 uVs 15:36:53 Integrated Voltage Measurement... 15:37:02 Voltage Offset Correction: v_off_ini = 366.140 uV, v_off_fin = 364.197 uV 15:37:02 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -495.323 uVs 15:37:02 Integrated Voltage Measurement... 15:37:11 Voltage Offset Correction: v_off_ini = 355.685 uV, v_off_fin = 367.079 uV 15:37:11 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 498.780 uVs 15:37:11 Integrated Voltage Measurement... 15:37:20 Voltage Offset Correction: v_off_ini = 367.973 uV, v_off_fin = 367.452 uV 15:37:20 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -496.959 uVs 15:37:20 Integrated Voltage Measurement... 15:37:29 Voltage Offset Correction: v_off_ini = 368.698 uV, v_off_fin = 377.685 uV 15:37:29 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 489.101 uVs 15:37:31 I2Y Average and RMS of Measurements VT = -496.447 +- 2.578 uVs I2Y = -45.674 +- 0.238 uTm^2 15:37:42 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 15:37:46 Integrated Voltage Measurement... 15:37:55 Voltage Offset Correction: v_off_ini = 391.399 uV, v_off_fin = 384.804 uV 15:37:55 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -83.117 uVs 15:37:55 Integrated Voltage Measurement... 15:38:04 Voltage Offset Correction: v_off_ini = 379.938 uV, v_off_fin = 398.409 uV 15:38:04 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 64.826 uVs 15:38:05 Integrated Voltage Measurement... 15:38:13 Voltage Offset Correction: v_off_ini = 380.187 uV, v_off_fin = 383.457 uV 15:38:14 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -60.818 uVs 15:38:14 Integrated Voltage Measurement... 15:38:22 Voltage Offset Correction: v_off_ini = 386.862 uV, v_off_fin = 386.151 uV 15:38:23 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 71.633 uVs 15:38:23 Integrated Voltage Measurement... 15:38:32 Voltage Offset Correction: v_off_ini = 389.963 uV, v_off_fin = 383.106 uV 15:38:32 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -68.614 uVs 15:38:32 Integrated Voltage Measurement... 15:38:41 Voltage Offset Correction: v_off_ini = 378.974 uV, v_off_fin = 378.286 uV 15:38:41 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 67.495 uVs 15:38:45 I1X Average and RMS of Measurements VT = -69.417 +- 3.306 uVs I1X = 13.884 +- 0.662 uTm 15:38:47 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 15:38:52 Integrated Voltage Measurement... 15:39:00 Voltage Offset Correction: v_off_ini = 380.016 uV, v_off_fin = 366.739 uV 15:39:01 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -178.869 uVs 15:39:01 Integrated Voltage Measurement... 15:39:09 Voltage Offset Correction: v_off_ini = 375.715 uV, v_off_fin = 390.388 uV 15:39:10 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 197.643 uVs 15:39:10 Integrated Voltage Measurement... 15:39:18 Voltage Offset Correction: v_off_ini = 366.184 uV, v_off_fin = 356.138 uV 15:39:19 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -180.371 uVs 15:39:19 Integrated Voltage Measurement... 15:39:27 Voltage Offset Correction: v_off_ini = 373.511 uV, v_off_fin = 390.689 uV 15:39:28 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 182.290 uVs 15:39:28 Integrated Voltage Measurement... 15:39:37 Voltage Offset Correction: v_off_ini = 365.537 uV, v_off_fin = 364.976 uV 15:39:37 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -185.609 uVs 15:39:37 Integrated Voltage Measurement... 15:39:45 Voltage Offset Correction: v_off_ini = 377.059 uV, v_off_fin = 380.378 uV 15:39:46 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 176.212 uVs 15:39:50 I1Y Average and RMS of Measurements VT = -183.499 +- 3.368 uVs I1Y = -36.697 +- 0.674 uTm 15:39:55 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 15:39:57 Integrated Voltage Measurement... 15:40:06 Voltage Offset Correction: v_off_ini = 374.149 uV, v_off_fin = 378.719 uV 15:40:07 Integrated Voltage Measurement dXYus = -1.0002 mm, dXYds = 0.0000 mm dXY = -1.0002 mm VT = -52.163 uVs 15:40:07 Integrated Voltage Measurement... 15:40:16 Voltage Offset Correction: v_off_ini = 397.559 uV, v_off_fin = 407.675 uV 15:40:16 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 10.944 uVs 15:40:16 Integrated Voltage Measurement... 15:40:25 Voltage Offset Correction: v_off_ini = 398.788 uV, v_off_fin = 376.058 uV 15:40:25 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -54.297 uVs 15:40:25 Integrated Voltage Measurement... 15:40:34 Voltage Offset Correction: v_off_ini = 387.277 uV, v_off_fin = 389.461 uV 15:40:34 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 17.557 uVs 15:40:34 Integrated Voltage Measurement... 15:40:43 Voltage Offset Correction: v_off_ini = 381.178 uV, v_off_fin = 363.651 uV 15:40:43 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -29.821 uVs 15:40:43 Integrated Voltage Measurement... 15:40:52 Voltage Offset Correction: v_off_ini = 382.280 uV, v_off_fin = 393.321 uV 15:40:52 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 52.824 uVs 15:40:54 I2X Average and RMS of Measurements VT = -36.267 +- 3.995 uVs I2X = 3.336 +- 0.368 uTm^2 15:40:59 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 15:41:02 Integrated Voltage Measurement... 15:41:11 Voltage Offset Correction: v_off_ini = 377.442 uV, v_off_fin = 377.471 uV 15:41:11 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -489.706 uVs 15:41:11 Integrated Voltage Measurement... 15:41:20 Voltage Offset Correction: v_off_ini = 386.245 uV, v_off_fin = 364.480 uV 15:41:20 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 490.902 uVs 15:41:20 Integrated Voltage Measurement... 15:41:29 Voltage Offset Correction: v_off_ini = 373.052 uV, v_off_fin = 370.106 uV 15:41:29 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -520.367 uVs 15:41:29 Integrated Voltage Measurement... 15:41:38 Voltage Offset Correction: v_off_ini = 373.571 uV, v_off_fin = 378.322 uV 15:41:38 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 483.716 uVs 15:41:38 Integrated Voltage Measurement... 15:41:47 Voltage Offset Correction: v_off_ini = 372.386 uV, v_off_fin = 367.436 uV 15:41:47 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -497.761 uVs 15:41:47 Integrated Voltage Measurement... 15:41:56 Voltage Offset Correction: v_off_ini = 371.364 uV, v_off_fin = 366.521 uV 15:41:56 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 484.134 uVs 15:41:58 I2Y Average and RMS of Measurements VT = -494.431 +- 5.388 uVs I2Y = -45.488 +- 0.496 uTm^2 15:42:09 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 15:42:13 Integrated Voltage Measurement... 15:42:22 Voltage Offset Correction: v_off_ini = 369.061 uV, v_off_fin = 372.088 uV 15:42:22 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -94.576 uVs 15:42:22 Integrated Voltage Measurement... 15:42:31 Voltage Offset Correction: v_off_ini = 358.943 uV, v_off_fin = 387.391 uV 15:42:31 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 70.014 uVs 15:42:31 Integrated Voltage Measurement... 15:42:40 Voltage Offset Correction: v_off_ini = 371.006 uV, v_off_fin = 359.247 uV 15:42:40 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -81.061 uVs 15:42:40 Integrated Voltage Measurement... 15:42:49 Voltage Offset Correction: v_off_ini = 365.365 uV, v_off_fin = 369.821 uV 15:42:49 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 89.460 uVs 15:42:49 Integrated Voltage Measurement... 15:42:58 Voltage Offset Correction: v_off_ini = 389.885 uV, v_off_fin = 376.864 uV 15:42:58 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -83.006 uVs 15:42:58 Integrated Voltage Measurement... 15:43:07 Voltage Offset Correction: v_off_ini = 360.475 uV, v_off_fin = 351.838 uV 15:43:07 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 86.646 uVs 15:43:12 I1X Average and RMS of Measurements VT = -84.127 +- 1.308 uVs I1X = 16.826 +- 0.261 uTm 15:43:14 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 15:43:18 Integrated Voltage Measurement... 15:43:27 Voltage Offset Correction: v_off_ini = 367.483 uV, v_off_fin = 359.773 uV 15:43:27 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -163.170 uVs 15:43:27 Integrated Voltage Measurement... 15:43:36 Voltage Offset Correction: v_off_ini = 390.210 uV, v_off_fin = 382.396 uV 15:43:36 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 197.717 uVs 15:43:36 Integrated Voltage Measurement... 15:43:45 Voltage Offset Correction: v_off_ini = 376.128 uV, v_off_fin = 363.958 uV 15:43:45 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -183.768 uVs 15:43:45 Integrated Voltage Measurement... 15:43:54 Voltage Offset Correction: v_off_ini = 374.837 uV, v_off_fin = 392.630 uV 15:43:54 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 181.419 uVs 15:43:54 Integrated Voltage Measurement... 15:44:03 Voltage Offset Correction: v_off_ini = 368.174 uV, v_off_fin = 359.880 uV 15:44:03 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -191.237 uVs 15:44:03 Integrated Voltage Measurement... 15:44:12 Voltage Offset Correction: v_off_ini = 380.278 uV, v_off_fin = 386.499 uV 15:44:12 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 188.707 uVs 15:44:17 I1Y Average and RMS of Measurements VT = -184.336 +- 4.080 uVs I1Y = -36.865 +- 0.816 uTm 15:44:22 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 15:44:24 Integrated Voltage Measurement... 15:44:33 Voltage Offset Correction: v_off_ini = 368.884 uV, v_off_fin = 376.686 uV 15:44:33 Integrated Voltage Measurement dXYus = -1.0002 mm, dXYds = 0.0000 mm dXY = -1.0002 mm VT = -41.285 uVs 15:44:33 Integrated Voltage Measurement... 15:44:42 Voltage Offset Correction: v_off_ini = 386.139 uV, v_off_fin = 384.377 uV 15:44:42 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 48.046 uVs 15:44:42 Integrated Voltage Measurement... 15:44:51 Voltage Offset Correction: v_off_ini = 372.073 uV, v_off_fin = 355.180 uV 15:44:51 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -57.187 uVs 15:44:51 Integrated Voltage Measurement... 15:45:00 Voltage Offset Correction: v_off_ini = 374.131 uV, v_off_fin = 393.452 uV 15:45:00 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 69.424 uVs 15:45:00 Integrated Voltage Measurement... 15:45:09 Voltage Offset Correction: v_off_ini = 377.284 uV, v_off_fin = 369.928 uV 15:45:10 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -56.591 uVs 15:45:10 Integrated Voltage Measurement... 15:45:19 Voltage Offset Correction: v_off_ini = 374.417 uV, v_off_fin = 378.229 uV 15:45:19 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 57.197 uVs 15:45:21 I2X Average and RMS of Measurements VT = -54.955 +- 7.732 uVs I2X = 5.056 +- 0.712 uTm^2 15:45:26 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 15:45:29 Integrated Voltage Measurement... 15:45:38 Voltage Offset Correction: v_off_ini = 385.018 uV, v_off_fin = 384.477 uV 15:45:38 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -478.113 uVs 15:45:38 Integrated Voltage Measurement... 15:45:47 Voltage Offset Correction: v_off_ini = 384.802 uV, v_off_fin = 388.002 uV 15:45:47 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 474.074 uVs 15:45:47 Integrated Voltage Measurement... 15:45:56 Voltage Offset Correction: v_off_ini = 378.396 uV, v_off_fin = 384.973 uV 15:45:56 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -479.192 uVs 15:45:56 Integrated Voltage Measurement... 15:46:05 Voltage Offset Correction: v_off_ini = 373.701 uV, v_off_fin = 383.899 uV 15:46:05 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 475.790 uVs 15:46:05 Integrated Voltage Measurement... 15:46:14 Voltage Offset Correction: v_off_ini = 376.394 uV, v_off_fin = 388.165 uV 15:46:14 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -487.307 uVs 15:46:14 Integrated Voltage Measurement... 15:46:23 Voltage Offset Correction: v_off_ini = 390.320 uV, v_off_fin = 391.318 uV 15:46:23 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 493.248 uVs 15:46:25 I2Y Average and RMS of Measurements VT = -481.287 +- 6.382 uVs I2Y = -44.278 +- 0.587 uTm^2 15:46:36 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 15:46:40 Integrated Voltage Measurement... 15:46:49 Voltage Offset Correction: v_off_ini = 406.525 uV, v_off_fin = 400.225 uV 15:46:49 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.4998 mm dXY = -0.4999 mm VT = -90.965 uVs 15:46:49 Integrated Voltage Measurement... 15:46:58 Voltage Offset Correction: v_off_ini = 410.889 uV, v_off_fin = 407.506 uV 15:46:58 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 58.459 uVs 15:46:58 Integrated Voltage Measurement... 15:47:07 Voltage Offset Correction: v_off_ini = 402.956 uV, v_off_fin = 406.839 uV 15:47:07 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -88.594 uVs 15:47:07 Integrated Voltage Measurement... 15:47:16 Voltage Offset Correction: v_off_ini = 416.405 uV, v_off_fin = 407.641 uV 15:47:16 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 81.306 uVs 15:47:16 Integrated Voltage Measurement... 15:47:25 Voltage Offset Correction: v_off_ini = 422.511 uV, v_off_fin = 415.391 uV 15:47:25 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -85.787 uVs 15:47:25 Integrated Voltage Measurement... 15:47:34 Voltage Offset Correction: v_off_ini = 408.890 uV, v_off_fin = 413.594 uV 15:47:34 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 82.207 uVs 15:47:39 I1X Average and RMS of Measurements VT = -81.220 +- 4.618 uVs I1X = 16.245 +- 0.923 uTm 15:47:41 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 15:47:45 Integrated Voltage Measurement... 15:47:54 Voltage Offset Correction: v_off_ini = 411.289 uV, v_off_fin = 403.028 uV 15:47:54 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -180.862 uVs 15:47:54 Integrated Voltage Measurement... 15:48:03 Voltage Offset Correction: v_off_ini = 426.140 uV, v_off_fin = 425.613 uV 15:48:03 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 195.428 uVs 15:48:03 Integrated Voltage Measurement... 15:48:12 Voltage Offset Correction: v_off_ini = 417.266 uV, v_off_fin = 400.774 uV 15:48:12 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -188.956 uVs 15:48:12 Integrated Voltage Measurement... 15:48:21 Voltage Offset Correction: v_off_ini = 411.328 uV, v_off_fin = 423.277 uV 15:48:21 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 196.734 uVs 15:48:21 Integrated Voltage Measurement... 15:48:30 Voltage Offset Correction: v_off_ini = 409.114 uV, v_off_fin = 412.760 uV 15:48:30 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -185.196 uVs 15:48:30 Integrated Voltage Measurement... 15:48:39 Voltage Offset Correction: v_off_ini = 413.332 uV, v_off_fin = 427.829 uV 15:48:39 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 174.644 uVs 15:48:44 I1Y Average and RMS of Measurements VT = -186.970 +- 5.342 uVs I1Y = -37.392 +- 1.068 uTm 15:48:49 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 15:48:51 Integrated Voltage Measurement... 15:49:00 Voltage Offset Correction: v_off_ini = 407.907 uV, v_off_fin = 404.113 uV 15:49:00 Integrated Voltage Measurement dXYus = -1.0001 mm, dXYds = 0.0000 mm dXY = -1.0001 mm VT = -65.843 uVs 15:49:00 Integrated Voltage Measurement... 15:49:09 Voltage Offset Correction: v_off_ini = 419.324 uV, v_off_fin = 444.714 uV 15:49:09 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 48.642 uVs 15:49:09 Integrated Voltage Measurement... 15:49:18 Voltage Offset Correction: v_off_ini = 437.481 uV, v_off_fin = 414.276 uV 15:49:18 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -77.018 uVs 15:49:18 Integrated Voltage Measurement... 15:49:27 Voltage Offset Correction: v_off_ini = 431.047 uV, v_off_fin = 437.093 uV 15:49:27 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 62.960 uVs 15:49:27 Integrated Voltage Measurement... 15:49:36 Voltage Offset Correction: v_off_ini = 438.749 uV, v_off_fin = 421.004 uV 15:49:36 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -58.078 uVs 15:49:36 Integrated Voltage Measurement... 15:49:45 Voltage Offset Correction: v_off_ini = 416.437 uV, v_off_fin = 436.568 uV 15:49:45 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 68.256 uVs 15:49:48 I2X Average and RMS of Measurements VT = -63.466 +- 5.208 uVs I2X = 5.839 +- 0.479 uTm^2 15:49:53 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 15:49:56 Integrated Voltage Measurement... 15:50:05 Voltage Offset Correction: v_off_ini = 427.499 uV, v_off_fin = 412.722 uV 15:50:05 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -472.281 uVs 15:50:05 Integrated Voltage Measurement... 15:50:14 Voltage Offset Correction: v_off_ini = 425.047 uV, v_off_fin = 431.122 uV 15:50:14 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 485.514 uVs 15:50:14 Integrated Voltage Measurement... 15:50:23 Voltage Offset Correction: v_off_ini = 424.934 uV, v_off_fin = 434.692 uV 15:50:23 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = -461.247 uVs 15:50:23 Integrated Voltage Measurement... 15:50:32 Voltage Offset Correction: v_off_ini = 443.285 uV, v_off_fin = 433.148 uV 15:50:32 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 471.251 uVs 15:50:32 Integrated Voltage Measurement... 15:50:41 Voltage Offset Correction: v_off_ini = 432.267 uV, v_off_fin = 431.911 uV 15:50:41 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -476.445 uVs 15:50:41 Integrated Voltage Measurement... 15:50:50 Voltage Offset Correction: v_off_ini = 435.440 uV, v_off_fin = 429.810 uV 15:50:50 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 462.579 uVs 15:50:52 I2Y Average and RMS of Measurements VT = -471.553 +- 5.362 uVs I2Y = -43.384 +- 0.493 uTm^2 15:51:03 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 15:51:07 Integrated Voltage Measurement... 15:51:16 Voltage Offset Correction: v_off_ini = 435.478 uV, v_off_fin = 427.012 uV 15:51:16 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -93.985 uVs 15:51:16 Integrated Voltage Measurement... 15:51:25 Voltage Offset Correction: v_off_ini = 429.030 uV, v_off_fin = 432.312 uV 15:51:25 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 84.761 uVs 15:51:25 Integrated Voltage Measurement... 15:51:34 Voltage Offset Correction: v_off_ini = 435.459 uV, v_off_fin = 434.317 uV 15:51:34 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -102.734 uVs 15:51:34 Integrated Voltage Measurement... 15:51:43 Voltage Offset Correction: v_off_ini = 443.208 uV, v_off_fin = 436.130 uV 15:51:43 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 103.498 uVs 15:51:43 Integrated Voltage Measurement... 15:51:52 Voltage Offset Correction: v_off_ini = 451.901 uV, v_off_fin = 449.006 uV 15:51:52 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -96.346 uVs 15:51:52 Integrated Voltage Measurement... 15:52:01 Voltage Offset Correction: v_off_ini = 443.433 uV, v_off_fin = 440.568 uV 15:52:01 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 80.368 uVs 15:52:05 I1X Average and RMS of Measurements VT = -93.615 +- 6.731 uVs I1X = 18.723 +- 1.346 uTm 15:52:08 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 15:52:12 Integrated Voltage Measurement... 15:52:21 Voltage Offset Correction: v_off_ini = 457.485 uV, v_off_fin = 430.424 uV 15:52:21 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -166.784 uVs 15:52:21 Integrated Voltage Measurement... 15:52:30 Voltage Offset Correction: v_off_ini = 446.725 uV, v_off_fin = 467.255 uV 15:52:30 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 181.781 uVs 15:52:30 Integrated Voltage Measurement... 15:52:39 Voltage Offset Correction: v_off_ini = 466.773 uV, v_off_fin = 447.458 uV 15:52:39 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -181.112 uVs 15:52:39 Integrated Voltage Measurement... 15:52:48 Voltage Offset Correction: v_off_ini = 464.668 uV, v_off_fin = 468.823 uV 15:52:48 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 202.164 uVs 15:52:48 Integrated Voltage Measurement... 15:52:57 Voltage Offset Correction: v_off_ini = 468.477 uV, v_off_fin = 445.442 uV 15:52:57 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -189.641 uVs 15:52:57 Integrated Voltage Measurement... 15:53:06 Voltage Offset Correction: v_off_ini = 447.415 uV, v_off_fin = 472.952 uV 15:53:06 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 226.397 uVs 15:53:10 I1Y Average and RMS of Measurements VT = -191.313 +- 13.775 uVs I1Y = -38.260 +- 2.754 uTm 15:53:16 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 15:53:18 Integrated Voltage Measurement... 15:53:27 Voltage Offset Correction: v_off_ini = 450.586 uV, v_off_fin = 472.432 uV 15:53:27 Integrated Voltage Measurement dXYus = -1.0002 mm, dXYds = 0.0000 mm dXY = -1.0002 mm VT = -40.484 uVs 15:53:27 Integrated Voltage Measurement... 15:53:36 Voltage Offset Correction: v_off_ini = 477.234 uV, v_off_fin = 483.340 uV 15:53:36 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 61.985 uVs 15:53:36 Integrated Voltage Measurement... 15:53:45 Voltage Offset Correction: v_off_ini = 482.366 uV, v_off_fin = 469.892 uV 15:53:45 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -74.834 uVs 15:53:45 Integrated Voltage Measurement... 15:53:54 Voltage Offset Correction: v_off_ini = 473.216 uV, v_off_fin = 486.427 uV 15:53:54 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 71.895 uVs 15:53:54 Integrated Voltage Measurement... 15:54:03 Voltage Offset Correction: v_off_ini = 483.723 uV, v_off_fin = 487.488 uV 15:54:03 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -65.134 uVs 15:54:03 Integrated Voltage Measurement... 15:54:12 Voltage Offset Correction: v_off_ini = 478.480 uV, v_off_fin = 494.681 uV 15:54:12 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 77.646 uVs 15:54:15 I2X Average and RMS of Measurements VT = -65.330 +- 10.000 uVs I2X = 6.010 +- 0.920 uTm^2 15:54:20 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 15:54:22 Integrated Voltage Measurement... 15:54:31 Voltage Offset Correction: v_off_ini = 497.825 uV, v_off_fin = 501.223 uV 15:54:32 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = -471.428 uVs 15:54:32 Integrated Voltage Measurement... 15:54:40 Voltage Offset Correction: v_off_ini = 498.769 uV, v_off_fin = 497.419 uV 15:54:41 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 471.634 uVs 15:54:41 Integrated Voltage Measurement... 15:54:50 Voltage Offset Correction: v_off_ini = 490.552 uV, v_off_fin = 496.021 uV 15:54:50 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -461.441 uVs 15:54:50 Integrated Voltage Measurement... 15:54:59 Voltage Offset Correction: v_off_ini = 504.008 uV, v_off_fin = 504.933 uV 15:54:59 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 456.998 uVs 15:54:59 Integrated Voltage Measurement... 15:55:08 Voltage Offset Correction: v_off_ini = 491.115 uV, v_off_fin = 490.538 uV 15:55:08 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -468.820 uVs 15:55:08 Integrated Voltage Measurement... 15:55:17 Voltage Offset Correction: v_off_ini = 506.046 uV, v_off_fin = 494.390 uV 15:55:17 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 468.965 uVs 15:55:19 I2Y Average and RMS of Measurements VT = -466.548 +- 5.292 uVs I2Y = -42.923 +- 0.488 uTm^2 15:55:30 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 15:55:34 Integrated Voltage Measurement... 15:55:43 Voltage Offset Correction: v_off_ini = 509.700 uV, v_off_fin = 516.891 uV 15:55:43 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -97.123 uVs 15:55:43 Integrated Voltage Measurement... 15:55:52 Voltage Offset Correction: v_off_ini = 511.896 uV, v_off_fin = 508.576 uV 15:55:52 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 86.049 uVs 15:55:52 Integrated Voltage Measurement... 15:56:01 Voltage Offset Correction: v_off_ini = 504.618 uV, v_off_fin = 510.920 uV 15:56:01 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -93.840 uVs 15:56:01 Integrated Voltage Measurement... 15:56:10 Voltage Offset Correction: v_off_ini = 517.137 uV, v_off_fin = 524.259 uV 15:56:10 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 104.314 uVs 15:56:10 Integrated Voltage Measurement... 15:56:19 Voltage Offset Correction: v_off_ini = 524.754 uV, v_off_fin = 517.704 uV 15:56:19 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -109.466 uVs 15:56:19 Integrated Voltage Measurement... 15:56:28 Voltage Offset Correction: v_off_ini = 528.471 uV, v_off_fin = 527.163 uV 15:56:28 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 85.703 uVs 15:56:32 I1X Average and RMS of Measurements VT = -96.082 +- 3.237 uVs I1X = 19.216 +- 0.647 uTm 15:56:35 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 15:56:39 Integrated Voltage Measurement... 15:56:48 Voltage Offset Correction: v_off_ini = 510.483 uV, v_off_fin = 504.498 uV 15:56:48 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -194.864 uVs 15:56:48 Integrated Voltage Measurement... 15:56:57 Voltage Offset Correction: v_off_ini = 505.350 uV, v_off_fin = 517.411 uV 15:56:57 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 209.119 uVs 15:56:57 Integrated Voltage Measurement... 15:57:06 Voltage Offset Correction: v_off_ini = 516.939 uV, v_off_fin = 502.847 uV 15:57:06 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -179.370 uVs 15:57:06 Integrated Voltage Measurement... 15:57:15 Voltage Offset Correction: v_off_ini = 519.126 uV, v_off_fin = 527.001 uV 15:57:15 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 193.963 uVs 15:57:15 Integrated Voltage Measurement... 15:57:24 Voltage Offset Correction: v_off_ini = 527.848 uV, v_off_fin = 512.610 uV 15:57:24 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -189.015 uVs 15:57:24 Integrated Voltage Measurement... 15:57:33 Voltage Offset Correction: v_off_ini = 525.636 uV, v_off_fin = 538.024 uV 15:57:33 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 216.631 uVs 15:57:37 I1Y Average and RMS of Measurements VT = -197.160 +- 7.428 uVs I1Y = -39.428 +- 1.485 uTm 15:57:42 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 15:57:45 Integrated Voltage Measurement... 15:57:54 Voltage Offset Correction: v_off_ini = 530.563 uV, v_off_fin = 527.748 uV 15:57:54 Integrated Voltage Measurement dXYus = -1.0001 mm, dXYds = 0.0000 mm dXY = -1.0001 mm VT = -75.138 uVs 15:57:54 Integrated Voltage Measurement... 15:58:03 Voltage Offset Correction: v_off_ini = 529.265 uV, v_off_fin = 558.123 uV 15:58:03 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 76.189 uVs 15:58:03 Integrated Voltage Measurement... 15:58:12 Voltage Offset Correction: v_off_ini = 537.033 uV, v_off_fin = 534.994 uV 15:58:12 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -74.160 uVs 15:58:12 Integrated Voltage Measurement... 15:58:21 Voltage Offset Correction: v_off_ini = 540.017 uV, v_off_fin = 552.288 uV 15:58:21 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 79.053 uVs 15:58:21 Integrated Voltage Measurement... 15:58:30 Voltage Offset Correction: v_off_ini = 537.651 uV, v_off_fin = 522.823 uV 15:58:30 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -85.885 uVs 15:58:30 Integrated Voltage Measurement... 15:58:39 Voltage Offset Correction: v_off_ini = 526.256 uV, v_off_fin = 535.117 uV 15:58:39 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 74.210 uVs 15:58:41 I2X Average and RMS of Measurements VT = -77.439 +- 1.884 uVs I2X = 7.125 +- 0.173 uTm^2 15:58:46 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 15:58:49 Integrated Voltage Measurement... 15:58:58 Voltage Offset Correction: v_off_ini = 505.871 uV, v_off_fin = 504.715 uV 15:58:58 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -460.518 uVs 15:58:58 Integrated Voltage Measurement... 15:59:07 Voltage Offset Correction: v_off_ini = 500.886 uV, v_off_fin = 506.837 uV 15:59:07 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 453.297 uVs 15:59:07 Integrated Voltage Measurement... 15:59:16 Voltage Offset Correction: v_off_ini = 509.974 uV, v_off_fin = 505.811 uV 15:59:16 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -456.206 uVs 15:59:16 Integrated Voltage Measurement... 15:59:25 Voltage Offset Correction: v_off_ini = 510.654 uV, v_off_fin = 507.121 uV 15:59:25 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 476.154 uVs 15:59:25 Integrated Voltage Measurement... 15:59:34 Voltage Offset Correction: v_off_ini = 506.744 uV, v_off_fin = 503.028 uV 15:59:34 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -470.106 uVs 15:59:34 Integrated Voltage Measurement... 15:59:43 Voltage Offset Correction: v_off_ini = 509.942 uV, v_off_fin = 515.053 uV 15:59:43 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 446.686 uVs 15:59:45 I2Y Average and RMS of Measurements VT = -460.494 +- 4.066 uVs I2Y = -42.365 +- 0.374 uTm^2 15:59:56 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:00:00 Integrated Voltage Measurement... 16:00:09 Voltage Offset Correction: v_off_ini = 497.848 uV, v_off_fin = 493.327 uV 16:00:09 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -113.623 uVs 16:00:09 Integrated Voltage Measurement... 16:00:18 Voltage Offset Correction: v_off_ini = 488.536 uV, v_off_fin = 501.592 uV 16:00:18 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = 105.844 uVs 16:00:18 Integrated Voltage Measurement... 16:00:27 Voltage Offset Correction: v_off_ini = 492.388 uV, v_off_fin = 484.600 uV 16:00:27 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -78.599 uVs 16:00:27 Integrated Voltage Measurement... 16:00:36 Voltage Offset Correction: v_off_ini = 491.805 uV, v_off_fin = 495.969 uV 16:00:36 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 107.788 uVs 16:00:36 Integrated Voltage Measurement... 16:00:45 Voltage Offset Correction: v_off_ini = 492.322 uV, v_off_fin = 486.191 uV 16:00:45 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -89.916 uVs 16:00:45 Integrated Voltage Measurement... 16:00:54 Voltage Offset Correction: v_off_ini = 483.920 uV, v_off_fin = 499.317 uV 16:00:54 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 96.584 uVs 16:00:59 I1X Average and RMS of Measurements VT = -98.726 +- 7.784 uVs I1X = 19.745 +- 1.557 uTm 16:01:01 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:01:05 Integrated Voltage Measurement... 16:01:14 Voltage Offset Correction: v_off_ini = 490.211 uV, v_off_fin = 481.048 uV 16:01:14 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -177.144 uVs 16:01:14 Integrated Voltage Measurement... 16:01:23 Voltage Offset Correction: v_off_ini = 491.851 uV, v_off_fin = 511.337 uV 16:01:23 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 210.876 uVs 16:01:23 Integrated Voltage Measurement... 16:01:32 Voltage Offset Correction: v_off_ini = 495.363 uV, v_off_fin = 474.697 uV 16:01:32 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -193.015 uVs 16:01:32 Integrated Voltage Measurement... 16:01:41 Voltage Offset Correction: v_off_ini = 479.576 uV, v_off_fin = 502.345 uV 16:01:41 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 197.174 uVs 16:01:41 Integrated Voltage Measurement... 16:01:50 Voltage Offset Correction: v_off_ini = 488.291 uV, v_off_fin = 480.146 uV 16:01:50 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -176.262 uVs 16:01:50 Integrated Voltage Measurement... 16:01:59 Voltage Offset Correction: v_off_ini = 488.104 uV, v_off_fin = 506.282 uV 16:01:59 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5002 mm dXY = -0.5001 mm VT = 203.490 uVs 16:02:03 I1Y Average and RMS of Measurements VT = -192.993 +- 2.249 uVs I1Y = -38.595 +- 0.450 uTm 16:02:09 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:02:11 Integrated Voltage Measurement... 16:02:20 Voltage Offset Correction: v_off_ini = 491.982 uV, v_off_fin = 490.102 uV 16:02:20 Integrated Voltage Measurement dXYus = -1.0002 mm, dXYds = 0.0000 mm dXY = -1.0002 mm VT = -95.149 uVs 16:02:20 Integrated Voltage Measurement... 16:02:29 Voltage Offset Correction: v_off_ini = 499.449 uV, v_off_fin = 530.315 uV 16:02:29 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 66.748 uVs 16:02:29 Integrated Voltage Measurement... 16:02:38 Voltage Offset Correction: v_off_ini = 510.814 uV, v_off_fin = 499.273 uV 16:02:38 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -105.225 uVs 16:02:38 Integrated Voltage Measurement... 16:02:47 Voltage Offset Correction: v_off_ini = 495.530 uV, v_off_fin = 513.557 uV 16:02:47 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 77.491 uVs 16:02:47 Integrated Voltage Measurement... 16:02:56 Voltage Offset Correction: v_off_ini = 501.143 uV, v_off_fin = 503.647 uV 16:02:56 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -100.972 uVs 16:02:56 Integrated Voltage Measurement... 16:03:05 Voltage Offset Correction: v_off_ini = 495.044 uV, v_off_fin = 505.228 uV 16:03:05 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 82.942 uVs 16:03:08 I2X Average and RMS of Measurements VT = -88.088 +- 5.054 uVs I2X = 8.104 +- 0.465 uTm^2 16:03:13 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:03:15 Integrated Voltage Measurement... 16:03:24 Voltage Offset Correction: v_off_ini = 504.152 uV, v_off_fin = 496.528 uV 16:03:24 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -460.868 uVs 16:03:24 Integrated Voltage Measurement... 16:03:33 Voltage Offset Correction: v_off_ini = 496.372 uV, v_off_fin = 489.295 uV 16:03:33 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0001 mm dXY = -1.0000 mm VT = 461.404 uVs 16:03:33 Integrated Voltage Measurement... 16:03:42 Voltage Offset Correction: v_off_ini = 516.314 uV, v_off_fin = 498.919 uV 16:03:42 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = -0.0001 mm dXY = 1.0000 mm VT = -470.756 uVs 16:03:42 Integrated Voltage Measurement... 16:03:51 Voltage Offset Correction: v_off_ini = 506.866 uV, v_off_fin = 500.741 uV 16:03:51 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 453.247 uVs 16:03:52 Integrated Voltage Measurement... 16:04:00 Voltage Offset Correction: v_off_ini = 500.668 uV, v_off_fin = 501.088 uV 16:04:01 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -465.728 uVs 16:04:01 Integrated Voltage Measurement... 16:04:09 Voltage Offset Correction: v_off_ini = 502.127 uV, v_off_fin = 501.092 uV 16:04:10 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 461.718 uVs 16:04:12 I2Y Average and RMS of Measurements VT = -462.287 +- 1.075 uVs I2Y = -42.530 +- 0.099 uTm^2 16:04:24 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:04:28 Integrated Voltage Measurement... 16:04:37 Voltage Offset Correction: v_off_ini = 501.279 uV, v_off_fin = 502.411 uV 16:04:37 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -94.478 uVs 16:04:37 Integrated Voltage Measurement... 16:04:46 Voltage Offset Correction: v_off_ini = 497.155 uV, v_off_fin = 496.636 uV 16:04:46 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 96.995 uVs 16:04:46 Integrated Voltage Measurement... 16:04:55 Voltage Offset Correction: v_off_ini = 509.781 uV, v_off_fin = 496.762 uV 16:04:55 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -107.616 uVs 16:04:55 Integrated Voltage Measurement... 16:05:04 Voltage Offset Correction: v_off_ini = 501.022 uV, v_off_fin = 490.933 uV 16:05:04 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 97.654 uVs 16:05:04 Integrated Voltage Measurement... 16:05:13 Voltage Offset Correction: v_off_ini = 493.337 uV, v_off_fin = 496.694 uV 16:05:13 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -86.823 uVs 16:05:13 Integrated Voltage Measurement... 16:05:22 Voltage Offset Correction: v_off_ini = 498.772 uV, v_off_fin = 508.692 uV 16:05:22 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 102.131 uVs 16:05:26 I1X Average and RMS of Measurements VT = -97.616 +- 3.586 uVs I1X = 19.523 +- 0.717 uTm 16:05:29 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:05:33 Integrated Voltage Measurement... 16:05:42 Voltage Offset Correction: v_off_ini = 512.045 uV, v_off_fin = 498.896 uV 16:05:42 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -189.787 uVs 16:05:42 Integrated Voltage Measurement... 16:05:51 Voltage Offset Correction: v_off_ini = 512.156 uV, v_off_fin = 538.496 uV 16:05:51 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 203.944 uVs 16:05:51 Integrated Voltage Measurement... 16:06:00 Voltage Offset Correction: v_off_ini = 525.203 uV, v_off_fin = 506.306 uV 16:06:00 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -198.491 uVs 16:06:00 Integrated Voltage Measurement... 16:06:09 Voltage Offset Correction: v_off_ini = 507.728 uV, v_off_fin = 526.285 uV 16:06:09 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 189.879 uVs 16:06:09 Integrated Voltage Measurement... 16:06:18 Voltage Offset Correction: v_off_ini = 512.415 uV, v_off_fin = 496.634 uV 16:06:18 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -206.460 uVs 16:06:18 Integrated Voltage Measurement... 16:06:27 Voltage Offset Correction: v_off_ini = 496.477 uV, v_off_fin = 513.217 uV 16:06:27 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 188.753 uVs 16:06:31 I1Y Average and RMS of Measurements VT = -196.219 +- 1.470 uVs I1Y = -39.241 +- 0.293 uTm 16:06:36 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:06:39 Integrated Voltage Measurement... 16:06:48 Voltage Offset Correction: v_off_ini = 506.302 uV, v_off_fin = 493.320 uV 16:06:48 Integrated Voltage Measurement dXYus = -1.0002 mm, dXYds = 0.0000 mm dXY = -1.0002 mm VT = -95.749 uVs 16:06:48 Integrated Voltage Measurement... 16:06:57 Voltage Offset Correction: v_off_ini = 489.435 uV, v_off_fin = 500.364 uV 16:06:57 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 81.043 uVs 16:06:57 Integrated Voltage Measurement... 16:07:06 Voltage Offset Correction: v_off_ini = 503.257 uV, v_off_fin = 491.417 uV 16:07:06 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -81.944 uVs 16:07:06 Integrated Voltage Measurement... 16:07:15 Voltage Offset Correction: v_off_ini = 501.207 uV, v_off_fin = 503.029 uV 16:07:15 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 96.414 uVs 16:07:15 Integrated Voltage Measurement... 16:07:24 Voltage Offset Correction: v_off_ini = 502.884 uV, v_off_fin = 494.182 uV 16:07:24 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -80.024 uVs 16:07:24 Integrated Voltage Measurement... 16:07:33 Voltage Offset Correction: v_off_ini = 498.039 uV, v_off_fin = 512.433 uV 16:07:33 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 93.968 uVs 16:07:35 I2X Average and RMS of Measurements VT = -88.190 +- 0.903 uVs I2X = 8.113 +- 0.083 uTm^2 16:07:40 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:07:43 Integrated Voltage Measurement... 16:07:52 Voltage Offset Correction: v_off_ini = 491.388 uV, v_off_fin = 505.556 uV 16:07:52 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -425.666 uVs 16:07:52 Integrated Voltage Measurement... 16:08:01 Voltage Offset Correction: v_off_ini = 510.687 uV, v_off_fin = 509.169 uV 16:08:01 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 429.849 uVs 16:08:01 Integrated Voltage Measurement... 16:08:10 Voltage Offset Correction: v_off_ini = 516.323 uV, v_off_fin = 512.581 uV 16:08:10 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -432.848 uVs 16:08:10 Integrated Voltage Measurement... 16:08:19 Voltage Offset Correction: v_off_ini = 512.268 uV, v_off_fin = 508.043 uV 16:08:19 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 436.662 uVs 16:08:19 Integrated Voltage Measurement... 16:08:28 Voltage Offset Correction: v_off_ini = 519.535 uV, v_off_fin = 509.069 uV 16:08:28 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -444.660 uVs 16:08:28 Integrated Voltage Measurement... 16:08:37 Voltage Offset Correction: v_off_ini = 508.942 uV, v_off_fin = 516.057 uV 16:08:37 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 448.629 uVs 16:08:39 I2Y Average and RMS of Measurements VT = -436.386 +- 7.796 uVs I2Y = -40.147 +- 0.717 uTm^2 16:08:51 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:08:56 Integrated Voltage Measurement... 16:09:05 Voltage Offset Correction: v_off_ini = 517.209 uV, v_off_fin = 529.354 uV 16:09:05 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -96.115 uVs 16:09:05 Integrated Voltage Measurement... 16:09:14 Voltage Offset Correction: v_off_ini = 539.539 uV, v_off_fin = 530.748 uV 16:09:14 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 111.674 uVs 16:09:14 Integrated Voltage Measurement... 16:09:23 Voltage Offset Correction: v_off_ini = 519.426 uV, v_off_fin = 533.612 uV 16:09:23 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -115.530 uVs 16:09:23 Integrated Voltage Measurement... 16:09:32 Voltage Offset Correction: v_off_ini = 545.551 uV, v_off_fin = 541.944 uV 16:09:32 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 91.654 uVs 16:09:32 Integrated Voltage Measurement... 16:09:41 Voltage Offset Correction: v_off_ini = 546.472 uV, v_off_fin = 542.026 uV 16:09:41 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -118.933 uVs 16:09:41 Integrated Voltage Measurement... 16:09:50 Voltage Offset Correction: v_off_ini = 556.720 uV, v_off_fin = 546.065 uV 16:09:50 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 95.825 uVs 16:09:54 I1X Average and RMS of Measurements VT = -104.955 +- 1.718 uVs I1X = 20.991 +- 0.344 uTm 16:09:56 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:10:01 Integrated Voltage Measurement... 16:10:09 Voltage Offset Correction: v_off_ini = 546.389 uV, v_off_fin = 541.138 uV 16:10:10 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -193.556 uVs 16:10:10 Integrated Voltage Measurement... 16:10:18 Voltage Offset Correction: v_off_ini = 559.720 uV, v_off_fin = 560.624 uV 16:10:18 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 191.001 uVs 16:10:19 Integrated Voltage Measurement... 16:10:27 Voltage Offset Correction: v_off_ini = 555.259 uV, v_off_fin = 542.517 uV 16:10:27 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -181.908 uVs 16:10:28 Integrated Voltage Measurement... 16:10:36 Voltage Offset Correction: v_off_ini = 564.255 uV, v_off_fin = 572.487 uV 16:10:36 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 196.793 uVs 16:10:37 Integrated Voltage Measurement... 16:10:45 Voltage Offset Correction: v_off_ini = 562.509 uV, v_off_fin = 531.379 uV 16:10:45 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -190.691 uVs 16:10:45 Integrated Voltage Measurement... 16:10:54 Voltage Offset Correction: v_off_ini = 540.145 uV, v_off_fin = 554.925 uV 16:10:54 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 204.045 uVs 16:10:59 I1Y Average and RMS of Measurements VT = -192.999 +- 3.312 uVs I1Y = -38.598 +- 0.663 uTm 16:11:04 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:11:06 Integrated Voltage Measurement... 16:11:15 Voltage Offset Correction: v_off_ini = 557.224 uV, v_off_fin = 563.686 uV 16:11:15 Integrated Voltage Measurement dXYus = -1.0002 mm, dXYds = 0.0000 mm dXY = -1.0002 mm VT = -118.323 uVs 16:11:15 Integrated Voltage Measurement... 16:11:24 Voltage Offset Correction: v_off_ini = 545.463 uV, v_off_fin = 569.825 uV 16:11:24 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 96.927 uVs 16:11:24 Integrated Voltage Measurement... 16:11:33 Voltage Offset Correction: v_off_ini = 550.227 uV, v_off_fin = 545.523 uV 16:11:33 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -93.889 uVs 16:11:33 Integrated Voltage Measurement... 16:11:42 Voltage Offset Correction: v_off_ini = 549.844 uV, v_off_fin = 557.659 uV 16:11:42 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 93.688 uVs 16:11:42 Integrated Voltage Measurement... 16:11:51 Voltage Offset Correction: v_off_ini = 552.922 uV, v_off_fin = 537.385 uV 16:11:51 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -75.711 uVs 16:11:52 Integrated Voltage Measurement... 16:12:00 Voltage Offset Correction: v_off_ini = 546.561 uV, v_off_fin = 551.072 uV 16:12:01 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 94.515 uVs 16:12:03 I2X Average and RMS of Measurements VT = -95.509 +- 9.271 uVs I2X = 8.787 +- 0.853 uTm^2 16:12:08 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:12:11 Integrated Voltage Measurement... 16:12:20 Voltage Offset Correction: v_off_ini = 545.234 uV, v_off_fin = 543.934 uV 16:12:20 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -422.248 uVs 16:12:20 Integrated Voltage Measurement... 16:12:29 Voltage Offset Correction: v_off_ini = 535.892 uV, v_off_fin = 546.605 uV 16:12:29 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 435.488 uVs 16:12:29 Integrated Voltage Measurement... 16:12:38 Voltage Offset Correction: v_off_ini = 538.875 uV, v_off_fin = 548.635 uV 16:12:38 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -430.567 uVs 16:12:38 Integrated Voltage Measurement... 16:12:47 Voltage Offset Correction: v_off_ini = 547.468 uV, v_off_fin = 551.745 uV 16:12:47 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 435.536 uVs 16:12:47 Integrated Voltage Measurement... 16:12:56 Voltage Offset Correction: v_off_ini = 543.411 uV, v_off_fin = 539.489 uV 16:12:56 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -422.945 uVs 16:12:56 Integrated Voltage Measurement... 16:13:05 Voltage Offset Correction: v_off_ini = 561.566 uV, v_off_fin = 537.492 uV 16:13:05 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 408.359 uVs 16:13:07 I2Y Average and RMS of Measurements VT = -425.857 +- 7.415 uVs I2Y = -39.179 +- 0.682 uTm^2 16:13:21 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:13:26 Integrated Voltage Measurement... 16:13:35 Voltage Offset Correction: v_off_ini = 549.683 uV, v_off_fin = 540.620 uV 16:13:35 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -126.224 uVs 16:13:35 Integrated Voltage Measurement... 16:13:44 Voltage Offset Correction: v_off_ini = 539.078 uV, v_off_fin = 555.269 uV 16:13:44 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 103.178 uVs 16:13:44 Integrated Voltage Measurement... 16:13:53 Voltage Offset Correction: v_off_ini = 544.225 uV, v_off_fin = 540.546 uV 16:13:53 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -126.837 uVs 16:13:53 Integrated Voltage Measurement... 16:14:02 Voltage Offset Correction: v_off_ini = 540.587 uV, v_off_fin = 533.724 uV 16:14:02 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 108.957 uVs 16:14:02 Integrated Voltage Measurement... 16:14:11 Voltage Offset Correction: v_off_ini = 541.477 uV, v_off_fin = 541.883 uV 16:14:11 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -118.014 uVs 16:14:11 Integrated Voltage Measurement... 16:14:20 Voltage Offset Correction: v_off_ini = 554.406 uV, v_off_fin = 540.872 uV 16:14:20 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = 110.834 uVs 16:14:24 I1X Average and RMS of Measurements VT = -115.674 +- 1.576 uVs I1X = 23.135 +- 0.315 uTm 16:14:26 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:14:31 Integrated Voltage Measurement... 16:14:40 Voltage Offset Correction: v_off_ini = 537.195 uV, v_off_fin = 530.593 uV 16:14:40 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -188.894 uVs 16:14:40 Integrated Voltage Measurement... 16:14:49 Voltage Offset Correction: v_off_ini = 541.856 uV, v_off_fin = 563.252 uV 16:14:49 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 211.077 uVs 16:14:49 Integrated Voltage Measurement... 16:14:58 Voltage Offset Correction: v_off_ini = 546.584 uV, v_off_fin = 537.330 uV 16:14:58 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -182.078 uVs 16:14:58 Integrated Voltage Measurement... 16:15:07 Voltage Offset Correction: v_off_ini = 538.237 uV, v_off_fin = 562.519 uV 16:15:07 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 200.378 uVs 16:15:07 Integrated Voltage Measurement... 16:15:16 Voltage Offset Correction: v_off_ini = 562.411 uV, v_off_fin = 548.879 uV 16:15:16 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -193.651 uVs 16:15:16 Integrated Voltage Measurement... 16:15:25 Voltage Offset Correction: v_off_ini = 547.210 uV, v_off_fin = 562.919 uV 16:15:25 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 212.444 uVs 16:15:29 I1Y Average and RMS of Measurements VT = -198.087 +- 5.008 uVs I1Y = -39.615 +- 1.003 uTm 16:15:34 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:15:37 Integrated Voltage Measurement... 16:15:46 Voltage Offset Correction: v_off_ini = 566.604 uV, v_off_fin = 568.454 uV 16:15:46 Integrated Voltage Measurement dXYus = -1.0002 mm, dXYds = 0.0000 mm dXY = -1.0002 mm VT = -95.936 uVs 16:15:46 Integrated Voltage Measurement... 16:15:55 Voltage Offset Correction: v_off_ini = 588.320 uV, v_off_fin = 588.508 uV 16:15:55 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 72.801 uVs 16:15:55 Integrated Voltage Measurement... 16:16:04 Voltage Offset Correction: v_off_ini = 585.503 uV, v_off_fin = 573.680 uV 16:16:04 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -87.572 uVs 16:16:04 Integrated Voltage Measurement... 16:16:13 Voltage Offset Correction: v_off_ini = 590.294 uV, v_off_fin = 601.126 uV 16:16:13 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 95.475 uVs 16:16:13 Integrated Voltage Measurement... 16:16:22 Voltage Offset Correction: v_off_ini = 589.854 uV, v_off_fin = 579.734 uV 16:16:22 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -98.580 uVs 16:16:22 Integrated Voltage Measurement... 16:16:31 Voltage Offset Correction: v_off_ini = 584.464 uV, v_off_fin = 593.628 uV 16:16:31 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 112.131 uVs 16:16:33 I2X Average and RMS of Measurements VT = -93.749 +- 8.711 uVs I2X = 8.625 +- 0.802 uTm^2 16:16:38 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:16:41 Integrated Voltage Measurement... 16:16:50 Voltage Offset Correction: v_off_ini = 577.317 uV, v_off_fin = 584.443 uV 16:16:50 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -396.031 uVs 16:16:50 Integrated Voltage Measurement... 16:16:59 Voltage Offset Correction: v_off_ini = 582.151 uV, v_off_fin = 573.646 uV 16:16:59 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 398.344 uVs 16:16:59 Integrated Voltage Measurement... 16:17:08 Voltage Offset Correction: v_off_ini = 573.791 uV, v_off_fin = 571.983 uV 16:17:08 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -391.721 uVs 16:17:08 Integrated Voltage Measurement... 16:17:17 Voltage Offset Correction: v_off_ini = 574.738 uV, v_off_fin = 576.384 uV 16:17:17 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 389.295 uVs 16:17:17 Integrated Voltage Measurement... 16:17:26 Voltage Offset Correction: v_off_ini = 568.919 uV, v_off_fin = 570.846 uV 16:17:26 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -410.556 uVs 16:17:26 Integrated Voltage Measurement... 16:17:35 Voltage Offset Correction: v_off_ini = 568.443 uV, v_off_fin = 558.061 uV 16:17:35 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 400.486 uVs 16:17:38 I2Y Average and RMS of Measurements VT = -397.739 +- 6.141 uVs I2Y = -36.592 +- 0.565 uTm^2 16:17:52 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:17:56 Integrated Voltage Measurement... 16:18:05 Voltage Offset Correction: v_off_ini = 567.738 uV, v_off_fin = 569.077 uV 16:18:05 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -127.917 uVs 16:18:05 Integrated Voltage Measurement... 16:18:14 Voltage Offset Correction: v_off_ini = 568.380 uV, v_off_fin = 573.292 uV 16:18:14 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = 105.461 uVs 16:18:14 Integrated Voltage Measurement... 16:18:23 Voltage Offset Correction: v_off_ini = 559.059 uV, v_off_fin = 564.387 uV 16:18:23 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = -112.259 uVs 16:18:23 Integrated Voltage Measurement... 16:18:32 Voltage Offset Correction: v_off_ini = 567.264 uV, v_off_fin = 560.636 uV 16:18:32 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 122.971 uVs 16:18:32 Integrated Voltage Measurement... 16:18:41 Voltage Offset Correction: v_off_ini = 576.868 uV, v_off_fin = 574.520 uV 16:18:41 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -161.586 uVs 16:18:41 Integrated Voltage Measurement... 16:18:50 Voltage Offset Correction: v_off_ini = 575.631 uV, v_off_fin = 578.097 uV 16:18:50 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 113.586 uVs 16:18:54 I1X Average and RMS of Measurements VT = -123.963 +- 9.640 uVs I1X = 24.792 +- 1.928 uTm 16:18:57 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:19:01 Integrated Voltage Measurement... 16:19:10 Voltage Offset Correction: v_off_ini = 571.716 uV, v_off_fin = 567.729 uV 16:19:10 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -176.689 uVs 16:19:10 Integrated Voltage Measurement... 16:19:19 Voltage Offset Correction: v_off_ini = 578.786 uV, v_off_fin = 584.084 uV 16:19:19 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 198.736 uVs 16:19:19 Integrated Voltage Measurement... 16:19:28 Voltage Offset Correction: v_off_ini = 567.446 uV, v_off_fin = 565.182 uV 16:19:28 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -183.585 uVs 16:19:28 Integrated Voltage Measurement... 16:19:37 Voltage Offset Correction: v_off_ini = 564.500 uV, v_off_fin = 567.473 uV 16:19:37 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 202.348 uVs 16:19:37 Integrated Voltage Measurement... 16:19:46 Voltage Offset Correction: v_off_ini = 570.500 uV, v_off_fin = 549.031 uV 16:19:46 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -184.862 uVs 16:19:46 Integrated Voltage Measurement... 16:19:55 Voltage Offset Correction: v_off_ini = 557.598 uV, v_off_fin = 580.277 uV 16:19:55 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 192.737 uVs 16:19:59 I1Y Average and RMS of Measurements VT = -189.826 +- 2.264 uVs I1Y = -37.962 +- 0.452 uTm 16:20:04 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:20:07 Integrated Voltage Measurement... 16:20:16 Voltage Offset Correction: v_off_ini = 577.525 uV, v_off_fin = 564.940 uV 16:20:16 Integrated Voltage Measurement dXYus = -1.0002 mm, dXYds = 0.0000 mm dXY = -1.0002 mm VT = -89.410 uVs 16:20:16 Integrated Voltage Measurement... 16:20:25 Voltage Offset Correction: v_off_ini = 580.045 uV, v_off_fin = 591.849 uV 16:20:25 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 98.463 uVs 16:20:25 Integrated Voltage Measurement... 16:20:34 Voltage Offset Correction: v_off_ini = 574.194 uV, v_off_fin = 571.548 uV 16:20:34 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -86.460 uVs 16:20:34 Integrated Voltage Measurement... 16:20:43 Voltage Offset Correction: v_off_ini = 581.348 uV, v_off_fin = 595.099 uV 16:20:43 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 91.419 uVs 16:20:43 Integrated Voltage Measurement... 16:20:52 Voltage Offset Correction: v_off_ini = 581.736 uV, v_off_fin = 580.136 uV 16:20:52 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -87.363 uVs 16:20:52 Integrated Voltage Measurement... 16:21:01 Voltage Offset Correction: v_off_ini = 589.839 uV, v_off_fin = 611.951 uV 16:21:01 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 83.529 uVs 16:21:04 I2X Average and RMS of Measurements VT = -89.441 +- 3.484 uVs I2X = 8.228 +- 0.320 uTm^2 16:21:09 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:21:11 Integrated Voltage Measurement... 16:21:20 Voltage Offset Correction: v_off_ini = 600.492 uV, v_off_fin = 603.130 uV 16:21:20 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -382.051 uVs 16:21:20 Integrated Voltage Measurement... 16:21:29 Voltage Offset Correction: v_off_ini = 600.568 uV, v_off_fin = 595.610 uV 16:21:29 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 383.990 uVs 16:21:29 Integrated Voltage Measurement... 16:21:38 Voltage Offset Correction: v_off_ini = 599.308 uV, v_off_fin = 599.969 uV 16:21:38 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -375.615 uVs 16:21:38 Integrated Voltage Measurement... 16:21:47 Voltage Offset Correction: v_off_ini = 599.181 uV, v_off_fin = 602.403 uV 16:21:47 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 367.183 uVs 16:21:47 Integrated Voltage Measurement... 16:21:56 Voltage Offset Correction: v_off_ini = 597.782 uV, v_off_fin = 599.245 uV 16:21:56 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -383.190 uVs 16:21:56 Integrated Voltage Measurement... 16:22:05 Voltage Offset Correction: v_off_ini = 599.574 uV, v_off_fin = 594.800 uV 16:22:05 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 373.822 uVs 16:22:08 I2Y Average and RMS of Measurements VT = -377.642 +- 4.784 uVs I2Y = -34.743 +- 0.440 uTm^2 16:22:22 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:22:26 Integrated Voltage Measurement... 16:22:35 Voltage Offset Correction: v_off_ini = 604.726 uV, v_off_fin = 608.597 uV 16:22:35 Integrated Voltage Measurement dXYus = -0.4998 mm, dXYds = -0.4999 mm dXY = -0.4999 mm VT = -114.042 uVs 16:22:35 Integrated Voltage Measurement... 16:22:44 Voltage Offset Correction: v_off_ini = 604.966 uV, v_off_fin = 597.817 uV 16:22:44 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 105.361 uVs 16:22:44 Integrated Voltage Measurement... 16:22:53 Voltage Offset Correction: v_off_ini = 607.270 uV, v_off_fin = 603.852 uV 16:22:53 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -128.071 uVs 16:22:53 Integrated Voltage Measurement... 16:23:02 Voltage Offset Correction: v_off_ini = 596.249 uV, v_off_fin = 610.722 uV 16:23:02 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 101.208 uVs 16:23:02 Integrated Voltage Measurement... 16:23:11 Voltage Offset Correction: v_off_ini = 606.726 uV, v_off_fin = 617.441 uV 16:23:11 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -107.753 uVs 16:23:11 Integrated Voltage Measurement... 16:23:20 Voltage Offset Correction: v_off_ini = 630.461 uV, v_off_fin = 624.680 uV 16:23:20 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = 93.128 uVs 16:23:25 I1X Average and RMS of Measurements VT = -108.260 +- 5.885 uVs I1X = 21.653 +- 1.178 uTm 16:23:27 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:23:31 Integrated Voltage Measurement... 16:23:40 Voltage Offset Correction: v_off_ini = 606.813 uV, v_off_fin = 597.262 uV 16:23:40 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = -171.030 uVs 16:23:40 Integrated Voltage Measurement... 16:23:49 Voltage Offset Correction: v_off_ini = 615.847 uV, v_off_fin = 622.024 uV 16:23:49 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 199.313 uVs 16:23:49 Integrated Voltage Measurement... 16:23:58 Voltage Offset Correction: v_off_ini = 598.769 uV, v_off_fin = 595.210 uV 16:23:58 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -194.087 uVs 16:23:58 Integrated Voltage Measurement... 16:24:07 Voltage Offset Correction: v_off_ini = 608.858 uV, v_off_fin = 605.287 uV 16:24:07 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 207.326 uVs 16:24:07 Integrated Voltage Measurement... 16:24:16 Voltage Offset Correction: v_off_ini = 601.768 uV, v_off_fin = 585.746 uV 16:24:16 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -185.065 uVs 16:24:16 Integrated Voltage Measurement... 16:24:25 Voltage Offset Correction: v_off_ini = 595.282 uV, v_off_fin = 602.490 uV 16:24:25 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 191.037 uVs 16:24:29 I1Y Average and RMS of Measurements VT = -191.310 +- 6.748 uVs I1Y = -38.259 +- 1.348 uTm 16:24:35 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:24:37 Integrated Voltage Measurement... 16:24:46 Voltage Offset Correction: v_off_ini = 598.607 uV, v_off_fin = 591.386 uV 16:24:46 Integrated Voltage Measurement dXYus = -1.0002 mm, dXYds = 0.0000 mm dXY = -1.0002 mm VT = -92.596 uVs 16:24:46 Integrated Voltage Measurement... 16:24:55 Voltage Offset Correction: v_off_ini = 603.384 uV, v_off_fin = 610.240 uV 16:24:55 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 78.741 uVs 16:24:55 Integrated Voltage Measurement... 16:25:04 Voltage Offset Correction: v_off_ini = 600.116 uV, v_off_fin = 589.475 uV 16:25:04 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -89.105 uVs 16:25:04 Integrated Voltage Measurement... 16:25:13 Voltage Offset Correction: v_off_ini = 590.848 uV, v_off_fin = 603.750 uV 16:25:13 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 100.652 uVs 16:25:13 Integrated Voltage Measurement... 16:25:22 Voltage Offset Correction: v_off_ini = 601.150 uV, v_off_fin = 582.754 uV 16:25:22 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -95.765 uVs 16:25:22 Integrated Voltage Measurement... 16:25:31 Voltage Offset Correction: v_off_ini = 590.892 uV, v_off_fin = 606.125 uV 16:25:31 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 98.966 uVs 16:25:34 I2X Average and RMS of Measurements VT = -92.638 +- 5.031 uVs I2X = 8.522 +- 0.463 uTm^2 16:25:39 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:25:42 Integrated Voltage Measurement... 16:25:51 Voltage Offset Correction: v_off_ini = 595.231 uV, v_off_fin = 579.285 uV 16:25:51 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = -347.063 uVs 16:25:51 Integrated Voltage Measurement... 16:26:00 Voltage Offset Correction: v_off_ini = 574.663 uV, v_off_fin = 585.834 uV 16:26:00 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 349.475 uVs 16:26:00 Integrated Voltage Measurement... 16:26:09 Voltage Offset Correction: v_off_ini = 569.535 uV, v_off_fin = 579.093 uV 16:26:09 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -354.070 uVs 16:26:09 Integrated Voltage Measurement... 16:26:18 Voltage Offset Correction: v_off_ini = 577.689 uV, v_off_fin = 583.974 uV 16:26:18 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 354.353 uVs 16:26:18 Integrated Voltage Measurement... 16:26:27 Voltage Offset Correction: v_off_ini = 577.029 uV, v_off_fin = 579.654 uV 16:26:27 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -346.002 uVs 16:26:27 Integrated Voltage Measurement... 16:26:36 Voltage Offset Correction: v_off_ini = 586.728 uV, v_off_fin = 585.328 uV 16:26:36 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 345.882 uVs 16:26:38 I2Y Average and RMS of Measurements VT = -349.474 +- 3.482 uVs I2Y = -32.152 +- 0.320 uTm^2 16:26:52 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:26:56 Integrated Voltage Measurement... 16:27:05 Voltage Offset Correction: v_off_ini = 576.400 uV, v_off_fin = 589.635 uV 16:27:05 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -109.134 uVs 16:27:05 Integrated Voltage Measurement... 16:27:14 Voltage Offset Correction: v_off_ini = 585.078 uV, v_off_fin = 584.004 uV 16:27:14 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 83.387 uVs 16:27:14 Integrated Voltage Measurement... 16:27:23 Voltage Offset Correction: v_off_ini = 588.137 uV, v_off_fin = 581.518 uV 16:27:23 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -110.753 uVs 16:27:23 Integrated Voltage Measurement... 16:27:32 Voltage Offset Correction: v_off_ini = 584.917 uV, v_off_fin = 575.337 uV 16:27:32 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 111.574 uVs 16:27:32 Integrated Voltage Measurement... 16:27:41 Voltage Offset Correction: v_off_ini = 578.277 uV, v_off_fin = 574.778 uV 16:27:41 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -120.836 uVs 16:27:41 Integrated Voltage Measurement... 16:27:50 Voltage Offset Correction: v_off_ini = 580.950 uV, v_off_fin = 585.327 uV 16:27:50 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 115.619 uVs 16:27:55 I1X Average and RMS of Measurements VT = -108.551 +- 9.156 uVs I1X = 21.710 +- 1.831 uTm 16:27:57 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:28:01 Integrated Voltage Measurement... 16:28:10 Voltage Offset Correction: v_off_ini = 583.684 uV, v_off_fin = 571.615 uV 16:28:10 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -196.238 uVs 16:28:10 Integrated Voltage Measurement... 16:28:19 Voltage Offset Correction: v_off_ini = 580.644 uV, v_off_fin = 591.833 uV 16:28:19 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 209.813 uVs 16:28:19 Integrated Voltage Measurement... 16:28:28 Voltage Offset Correction: v_off_ini = 588.337 uV, v_off_fin = 577.445 uV 16:28:28 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -170.894 uVs 16:28:28 Integrated Voltage Measurement... 16:28:37 Voltage Offset Correction: v_off_ini = 585.029 uV, v_off_fin = 603.355 uV 16:28:37 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 194.672 uVs 16:28:37 Integrated Voltage Measurement... 16:28:46 Voltage Offset Correction: v_off_ini = 605.847 uV, v_off_fin = 594.241 uV 16:28:46 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -181.144 uVs 16:28:46 Integrated Voltage Measurement... 16:28:55 Voltage Offset Correction: v_off_ini = 602.210 uV, v_off_fin = 613.649 uV 16:28:55 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 193.900 uVs 16:29:00 I1Y Average and RMS of Measurements VT = -191.110 +- 8.645 uVs I1Y = -38.220 +- 1.729 uTm 16:29:05 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:29:07 Integrated Voltage Measurement... 16:29:16 Voltage Offset Correction: v_off_ini = 596.557 uV, v_off_fin = 595.281 uV 16:29:16 Integrated Voltage Measurement dXYus = -1.0002 mm, dXYds = 0.0000 mm dXY = -1.0002 mm VT = -88.536 uVs 16:29:16 Integrated Voltage Measurement... 16:29:25 Voltage Offset Correction: v_off_ini = 609.691 uV, v_off_fin = 614.557 uV 16:29:25 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 77.665 uVs 16:29:25 Integrated Voltage Measurement... 16:29:34 Voltage Offset Correction: v_off_ini = 601.564 uV, v_off_fin = 589.908 uV 16:29:34 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -85.255 uVs 16:29:34 Integrated Voltage Measurement... 16:29:43 Voltage Offset Correction: v_off_ini = 602.827 uV, v_off_fin = 622.721 uV 16:29:43 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 71.015 uVs 16:29:43 Integrated Voltage Measurement... 16:29:52 Voltage Offset Correction: v_off_ini = 606.688 uV, v_off_fin = 603.603 uV 16:29:52 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -94.198 uVs 16:29:52 Integrated Voltage Measurement... 16:30:01 Voltage Offset Correction: v_off_ini = 592.802 uV, v_off_fin = 609.729 uV 16:30:01 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 98.028 uVs 16:30:04 I2X Average and RMS of Measurements VT = -85.783 +- 7.581 uVs I2X = 7.892 +- 0.698 uTm^2 16:30:09 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:30:11 Integrated Voltage Measurement... 16:30:20 Voltage Offset Correction: v_off_ini = 625.724 uV, v_off_fin = 612.931 uV 16:30:20 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -339.975 uVs 16:30:21 Integrated Voltage Measurement... 16:30:29 Voltage Offset Correction: v_off_ini = 603.472 uV, v_off_fin = 606.834 uV 16:30:30 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 330.358 uVs 16:30:30 Integrated Voltage Measurement... 16:30:39 Voltage Offset Correction: v_off_ini = 607.136 uV, v_off_fin = 605.238 uV 16:30:39 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -343.392 uVs 16:30:39 Integrated Voltage Measurement... 16:30:48 Voltage Offset Correction: v_off_ini = 602.823 uV, v_off_fin = 603.334 uV 16:30:48 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 332.502 uVs 16:30:48 Integrated Voltage Measurement... 16:30:57 Voltage Offset Correction: v_off_ini = 599.587 uV, v_off_fin = 592.692 uV 16:30:57 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -345.390 uVs 16:30:57 Integrated Voltage Measurement... 16:31:06 Voltage Offset Correction: v_off_ini = 602.230 uV, v_off_fin = 601.032 uV 16:31:06 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 327.283 uVs 16:31:08 I2Y Average and RMS of Measurements VT = -336.483 +- 1.140 uVs I2Y = -30.956 +- 0.105 uTm^2 16:31:22 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:31:26 Integrated Voltage Measurement... 16:31:35 Voltage Offset Correction: v_off_ini = 600.781 uV, v_off_fin = 602.688 uV 16:31:35 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -105.371 uVs 16:31:35 Integrated Voltage Measurement... 16:31:44 Voltage Offset Correction: v_off_ini = 594.241 uV, v_off_fin = 613.081 uV 16:31:44 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 103.860 uVs 16:31:44 Integrated Voltage Measurement... 16:31:53 Voltage Offset Correction: v_off_ini = 590.097 uV, v_off_fin = 600.163 uV 16:31:53 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -92.229 uVs 16:31:53 Integrated Voltage Measurement... 16:32:02 Voltage Offset Correction: v_off_ini = 587.950 uV, v_off_fin = 596.927 uV 16:32:02 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 109.465 uVs 16:32:02 Integrated Voltage Measurement... 16:32:11 Voltage Offset Correction: v_off_ini = 589.637 uV, v_off_fin = 594.937 uV 16:32:11 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -111.520 uVs 16:32:11 Integrated Voltage Measurement... 16:32:20 Voltage Offset Correction: v_off_ini = 593.233 uV, v_off_fin = 587.098 uV 16:32:20 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 94.688 uVs 16:32:25 I1X Average and RMS of Measurements VT = -102.856 +- 1.549 uVs I1X = 20.571 +- 0.310 uTm 16:32:27 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:32:31 Integrated Voltage Measurement... 16:32:40 Voltage Offset Correction: v_off_ini = 577.287 uV, v_off_fin = 570.173 uV 16:32:40 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -182.246 uVs 16:32:40 Integrated Voltage Measurement... 16:32:49 Voltage Offset Correction: v_off_ini = 589.200 uV, v_off_fin = 601.718 uV 16:32:49 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 216.282 uVs 16:32:49 Integrated Voltage Measurement... 16:32:58 Voltage Offset Correction: v_off_ini = 584.550 uV, v_off_fin = 581.682 uV 16:32:58 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -189.987 uVs 16:32:58 Integrated Voltage Measurement... 16:33:07 Voltage Offset Correction: v_off_ini = 590.366 uV, v_off_fin = 598.573 uV 16:33:07 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 201.418 uVs 16:33:07 Integrated Voltage Measurement... 16:33:16 Voltage Offset Correction: v_off_ini = 598.809 uV, v_off_fin = 587.201 uV 16:33:16 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -181.282 uVs 16:33:16 Integrated Voltage Measurement... 16:33:25 Voltage Offset Correction: v_off_ini = 605.658 uV, v_off_fin = 598.239 uV 16:33:25 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 183.978 uVs 16:33:29 I1Y Average and RMS of Measurements VT = -192.532 +- 7.151 uVs I1Y = -38.505 +- 1.429 uTm 16:33:34 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:33:37 Integrated Voltage Measurement... 16:33:46 Voltage Offset Correction: v_off_ini = 576.650 uV, v_off_fin = 580.257 uV 16:33:46 Integrated Voltage Measurement dXYus = -1.0002 mm, dXYds = 0.0000 mm dXY = -1.0002 mm VT = -79.576 uVs 16:33:46 Integrated Voltage Measurement... 16:33:55 Voltage Offset Correction: v_off_ini = 576.797 uV, v_off_fin = 589.457 uV 16:33:55 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 86.399 uVs 16:33:55 Integrated Voltage Measurement... 16:34:04 Voltage Offset Correction: v_off_ini = 585.239 uV, v_off_fin = 569.519 uV 16:34:04 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -98.433 uVs 16:34:04 Integrated Voltage Measurement... 16:34:13 Voltage Offset Correction: v_off_ini = 575.015 uV, v_off_fin = 595.004 uV 16:34:13 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 83.483 uVs 16:34:13 Integrated Voltage Measurement... 16:34:22 Voltage Offset Correction: v_off_ini = 580.372 uV, v_off_fin = 578.535 uV 16:34:22 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -58.112 uVs 16:34:22 Integrated Voltage Measurement... 16:34:31 Voltage Offset Correction: v_off_ini = 591.993 uV, v_off_fin = 593.810 uV 16:34:31 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 76.375 uVs 16:34:34 I2X Average and RMS of Measurements VT = -80.396 +- 9.853 uVs I2X = 7.396 +- 0.906 uTm^2 16:34:39 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:34:41 Integrated Voltage Measurement... 16:34:50 Voltage Offset Correction: v_off_ini = 587.018 uV, v_off_fin = 581.858 uV 16:34:50 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -337.723 uVs 16:34:50 Integrated Voltage Measurement... 16:34:59 Voltage Offset Correction: v_off_ini = 574.957 uV, v_off_fin = 584.092 uV 16:34:59 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 334.851 uVs 16:34:59 Integrated Voltage Measurement... 16:35:08 Voltage Offset Correction: v_off_ini = 583.102 uV, v_off_fin = 580.318 uV 16:35:08 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -324.274 uVs 16:35:08 Integrated Voltage Measurement... 16:35:17 Voltage Offset Correction: v_off_ini = 589.364 uV, v_off_fin = 587.692 uV 16:35:17 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 314.151 uVs 16:35:17 Integrated Voltage Measurement... 16:35:26 Voltage Offset Correction: v_off_ini = 588.543 uV, v_off_fin = 593.868 uV 16:35:26 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -321.568 uVs 16:35:26 Integrated Voltage Measurement... 16:35:35 Voltage Offset Correction: v_off_ini = 595.374 uV, v_off_fin = 601.988 uV 16:35:35 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 311.928 uVs 16:35:38 I2Y Average and RMS of Measurements VT = -324.082 +- 8.688 uVs I2Y = -29.816 +- 0.799 uTm^2 16:35:52 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:35:56 Integrated Voltage Measurement... 16:36:05 Voltage Offset Correction: v_off_ini = 591.335 uV, v_off_fin = 585.056 uV 16:36:05 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.4998 mm dXY = -0.4999 mm VT = -102.909 uVs 16:36:05 Integrated Voltage Measurement... 16:36:14 Voltage Offset Correction: v_off_ini = 588.448 uV, v_off_fin = 588.365 uV 16:36:14 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 101.561 uVs 16:36:14 Integrated Voltage Measurement... 16:36:23 Voltage Offset Correction: v_off_ini = 586.578 uV, v_off_fin = 576.727 uV 16:36:23 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -91.224 uVs 16:36:23 Integrated Voltage Measurement... 16:36:32 Voltage Offset Correction: v_off_ini = 566.444 uV, v_off_fin = 572.562 uV 16:36:32 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 90.516 uVs 16:36:32 Integrated Voltage Measurement... 16:36:41 Voltage Offset Correction: v_off_ini = 567.194 uV, v_off_fin = 571.948 uV 16:36:41 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -90.439 uVs 16:36:41 Integrated Voltage Measurement... 16:36:50 Voltage Offset Correction: v_off_ini = 579.114 uV, v_off_fin = 583.699 uV 16:36:50 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 90.380 uVs 16:36:54 I1X Average and RMS of Measurements VT = -94.505 +- 5.469 uVs I1X = 18.903 +- 1.095 uTm 16:36:57 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:37:01 Integrated Voltage Measurement... 16:37:10 Voltage Offset Correction: v_off_ini = 577.004 uV, v_off_fin = 573.168 uV 16:37:10 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -174.873 uVs 16:37:10 Integrated Voltage Measurement... 16:37:19 Voltage Offset Correction: v_off_ini = 584.218 uV, v_off_fin = 601.407 uV 16:37:19 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 205.618 uVs 16:37:19 Integrated Voltage Measurement... 16:37:28 Voltage Offset Correction: v_off_ini = 601.840 uV, v_off_fin = 588.374 uV 16:37:28 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -189.891 uVs 16:37:28 Integrated Voltage Measurement... 16:37:37 Voltage Offset Correction: v_off_ini = 589.153 uV, v_off_fin = 597.252 uV 16:37:37 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 193.512 uVs 16:37:37 Integrated Voltage Measurement... 16:37:46 Voltage Offset Correction: v_off_ini = 586.068 uV, v_off_fin = 575.826 uV 16:37:46 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -194.656 uVs 16:37:46 Integrated Voltage Measurement... 16:37:55 Voltage Offset Correction: v_off_ini = 594.558 uV, v_off_fin = 607.782 uV 16:37:55 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 193.164 uVs 16:37:59 I1Y Average and RMS of Measurements VT = -191.952 +- 1.507 uVs I1Y = -38.390 +- 0.302 uTm 16:38:04 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:38:07 Integrated Voltage Measurement... 16:38:16 Voltage Offset Correction: v_off_ini = 591.075 uV, v_off_fin = 601.571 uV 16:38:16 Integrated Voltage Measurement dXYus = -1.0002 mm, dXYds = 0.0000 mm dXY = -1.0002 mm VT = -70.067 uVs 16:38:16 Integrated Voltage Measurement... 16:38:25 Voltage Offset Correction: v_off_ini = 594.104 uV, v_off_fin = 612.731 uV 16:38:25 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 94.916 uVs 16:38:25 Integrated Voltage Measurement... 16:38:34 Voltage Offset Correction: v_off_ini = 604.284 uV, v_off_fin = 587.725 uV 16:38:34 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -78.019 uVs 16:38:34 Integrated Voltage Measurement... 16:38:43 Voltage Offset Correction: v_off_ini = 600.638 uV, v_off_fin = 612.287 uV 16:38:43 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 78.564 uVs 16:38:43 Integrated Voltage Measurement... 16:38:52 Voltage Offset Correction: v_off_ini = 593.970 uV, v_off_fin = 593.650 uV 16:38:52 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -79.706 uVs 16:38:52 Integrated Voltage Measurement... 16:39:01 Voltage Offset Correction: v_off_ini = 600.991 uV, v_off_fin = 612.909 uV 16:39:01 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 83.512 uVs 16:39:04 I2X Average and RMS of Measurements VT = -80.797 +- 1.808 uVs I2X = 7.433 +- 0.166 uTm^2 16:39:09 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:39:11 Integrated Voltage Measurement... 16:39:20 Voltage Offset Correction: v_off_ini = 618.082 uV, v_off_fin = 617.323 uV 16:39:20 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -300.635 uVs 16:39:20 Integrated Voltage Measurement... 16:39:29 Voltage Offset Correction: v_off_ini = 617.737 uV, v_off_fin = 610.507 uV 16:39:29 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 317.064 uVs 16:39:29 Integrated Voltage Measurement... 16:39:38 Voltage Offset Correction: v_off_ini = 604.750 uV, v_off_fin = 605.468 uV 16:39:38 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -304.488 uVs 16:39:38 Integrated Voltage Measurement... 16:39:47 Voltage Offset Correction: v_off_ini = 609.517 uV, v_off_fin = 615.321 uV 16:39:47 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 292.017 uVs 16:39:47 Integrated Voltage Measurement... 16:39:56 Voltage Offset Correction: v_off_ini = 615.902 uV, v_off_fin = 616.901 uV 16:39:56 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -293.609 uVs 16:39:56 Integrated Voltage Measurement... 16:40:05 Voltage Offset Correction: v_off_ini = 614.872 uV, v_off_fin = 608.068 uV 16:40:05 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 313.828 uVs 16:40:08 I2Y Average and RMS of Measurements VT = -303.607 +- 4.327 uVs I2Y = -27.932 +- 0.398 uTm^2 16:40:10 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm