SLAC Magnetic Measurements Date: 07-17-2025 Time: 14:08:39 Project: LCLS-II-HE Magnet Type: Phase Shifter Magnet Name: HE_SXPS_16353 Dataset Number: 1 Run Purpose: Tuning Run Type: Stretched Wire Run Number: 5 Measurement Device: Operator: YL Comment: 14:08:48 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 14:08:50 Motion Parameters, X Controller, Axis 1 acc = 0.001000 m/s/s dcl = 0.001000 m/s/s vel = 0.001000 m/s pos = -0.000000 m kp = 188.600000 ki = 820.000000 kd = 491.400000 db = 0.000000 14:08:51 Motion Parameters, X Controller, Axis 2 acc = 0.001000 m/s/s dcl = 0.001000 m/s/s vel = 0.001000 m/s pos = 0.000000 m kp = 188.600000 ki = 820.000000 kd = 491.400000 db = 0.000000 14:08:53 Motion Parameters, Y Controller, Axis 1 acc = 0.001000 m/s/s dcl = 0.001000 m/s/s vel = 0.001000 m/s pos = 0.000000 m kp = 188.600000 ki = 820.000000 kd = 491.400000 db = 0.000000 14:08:55 Motion Parameters, Y Controller, Axis 2 acc = 0.001000 m/s/s dcl = 0.001000 m/s/s vel = 0.001000 m/s pos = 0.000000 m kp = 188.600000 ki = 820.000000 kd = 491.400000 db = 0.000000 14:09:05 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 14:09:09 Integrated Voltage Measurement... 14:09:18 Voltage Offset Correction: v_off_ini = 185.872 uV, v_off_fin = 176.590 uV 14:09:18 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -12.326 uVs 14:09:18 Integrated Voltage Measurement... 14:09:27 Voltage Offset Correction: v_off_ini = 201.385 uV, v_off_fin = 195.608 uV 14:09:27 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 14.712 uVs 14:09:27 Integrated Voltage Measurement... 14:09:36 Voltage Offset Correction: v_off_ini = 188.812 uV, v_off_fin = 188.239 uV 14:09:36 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -41.152 uVs 14:09:36 Integrated Voltage Measurement... 14:09:45 Voltage Offset Correction: v_off_ini = 219.262 uV, v_off_fin = 224.238 uV 14:09:45 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 34.162 uVs 14:09:45 Integrated Voltage Measurement... 14:09:54 Voltage Offset Correction: v_off_ini = 228.276 uV, v_off_fin = 223.007 uV 14:09:54 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -19.959 uVs 14:09:54 Integrated Voltage Measurement... 14:10:03 Voltage Offset Correction: v_off_ini = 219.937 uV, v_off_fin = 220.887 uV 14:10:03 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 4.024 uVs 14:10:07 I1X Average and RMS of Measurements VT = -21.056 +- 11.756 uVs I1X = 4.211 +- 2.351 uTm 14:10:10 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 14:10:14 Integrated Voltage Measurement... 14:10:23 Voltage Offset Correction: v_off_ini = 226.905 uV, v_off_fin = 227.435 uV 14:10:23 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.4998 mm dXY = 0.4999 mm VT = -184.141 uVs 14:10:23 Integrated Voltage Measurement... 14:10:32 Voltage Offset Correction: v_off_ini = 228.779 uV, v_off_fin = 238.267 uV 14:10:32 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 208.924 uVs 14:10:32 Integrated Voltage Measurement... 14:10:41 Voltage Offset Correction: v_off_ini = 236.485 uV, v_off_fin = 246.104 uV 14:10:41 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -183.650 uVs 14:10:41 Integrated Voltage Measurement... 14:10:50 Voltage Offset Correction: v_off_ini = 238.083 uV, v_off_fin = 263.278 uV 14:10:50 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 201.793 uVs 14:10:50 Integrated Voltage Measurement... 14:10:59 Voltage Offset Correction: v_off_ini = 251.922 uV, v_off_fin = 236.771 uV 14:10:59 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -195.569 uVs 14:10:59 Integrated Voltage Measurement... 14:11:08 Voltage Offset Correction: v_off_ini = 232.182 uV, v_off_fin = 247.014 uV 14:11:08 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 191.614 uVs 14:11:12 I1Y Average and RMS of Measurements VT = -194.282 +- 1.631 uVs I1Y = -38.857 +- 0.328 uTm 14:11:17 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 14:11:20 Integrated Voltage Measurement... 14:11:29 Voltage Offset Correction: v_off_ini = 231.318 uV, v_off_fin = 240.861 uV 14:11:29 Integrated Voltage Measurement dXYus = -1.0003 mm, dXYds = 0.0000 mm dXY = -1.0003 mm VT = 2.005 uVs 14:11:29 Integrated Voltage Measurement... 14:11:38 Voltage Offset Correction: v_off_ini = 236.813 uV, v_off_fin = 258.238 uV 14:11:38 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = -20.693 uVs 14:11:38 Integrated Voltage Measurement... 14:11:47 Voltage Offset Correction: v_off_ini = 233.640 uV, v_off_fin = 222.812 uV 14:11:47 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = 20.795 uVs 14:11:47 Integrated Voltage Measurement... 14:11:56 Voltage Offset Correction: v_off_ini = 247.858 uV, v_off_fin = 258.516 uV 14:11:56 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -7.654 uVs 14:11:56 Integrated Voltage Measurement... 14:12:05 Voltage Offset Correction: v_off_ini = 256.132 uV, v_off_fin = 241.413 uV 14:12:05 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 8.164 uVs 14:12:05 Integrated Voltage Measurement... 14:12:14 Voltage Offset Correction: v_off_ini = 253.747 uV, v_off_fin = 253.434 uV 14:12:14 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -18.483 uVs 14:12:17 I2X Average and RMS of Measurements VT = 12.966 +- 1.201 uVs I2X = -1.193 +- 0.110 uTm^2 14:12:22 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 14:12:24 Integrated Voltage Measurement... 14:12:33 Voltage Offset Correction: v_off_ini = 264.820 uV, v_off_fin = 275.094 uV 14:12:33 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -525.235 uVs 14:12:33 Integrated Voltage Measurement... 14:12:42 Voltage Offset Correction: v_off_ini = 267.232 uV, v_off_fin = 283.944 uV 14:12:43 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 541.971 uVs 14:12:43 Integrated Voltage Measurement... 14:12:52 Voltage Offset Correction: v_off_ini = 270.439 uV, v_off_fin = 260.837 uV 14:12:52 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -514.597 uVs 14:12:52 Integrated Voltage Measurement... 14:13:01 Voltage Offset Correction: v_off_ini = 261.121 uV, v_off_fin = 262.068 uV 14:13:01 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 533.673 uVs 14:13:01 Integrated Voltage Measurement... 14:13:10 Voltage Offset Correction: v_off_ini = 266.021 uV, v_off_fin = 267.089 uV 14:13:10 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -521.922 uVs 14:13:10 Integrated Voltage Measurement... 14:13:19 Voltage Offset Correction: v_off_ini = 262.112 uV, v_off_fin = 264.370 uV 14:13:19 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 525.578 uVs 14:13:21 I2Y Average and RMS of Measurements VT = -527.163 +- 4.557 uVs I2Y = -48.499 +- 0.419 uTm^2 14:13:32 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 14:13:36 Integrated Voltage Measurement... 14:13:45 Voltage Offset Correction: v_off_ini = 278.411 uV, v_off_fin = 261.801 uV 14:13:45 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -50.212 uVs 14:13:45 Integrated Voltage Measurement... 14:13:54 Voltage Offset Correction: v_off_ini = 245.969 uV, v_off_fin = 229.427 uV 14:13:54 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 50.350 uVs 14:13:54 Integrated Voltage Measurement... 14:14:03 Voltage Offset Correction: v_off_ini = 299.273 uV, v_off_fin = 331.347 uV 14:14:03 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.4999 mm dXY = -0.4999 mm VT = -186.923 uVs 14:14:03 Integrated Voltage Measurement... 14:14:12 Voltage Offset Correction: v_off_ini = 176.082 uV, v_off_fin = 194.425 uV 14:14:12 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 366.614 uVs 14:14:12 Integrated Voltage Measurement... 14:14:21 Voltage Offset Correction: v_off_ini = 252.844 uV, v_off_fin = 251.696 uV 14:14:21 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -31.381 uVs 14:14:21 Integrated Voltage Measurement... 14:14:30 Voltage Offset Correction: v_off_ini = 242.999 uV, v_off_fin = 234.619 uV 14:14:30 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 44.629 uVs 14:14:34 I1X Average and RMS of Measurements VT = -121.685 +- 109.775 uVs I1X = 24.338 +- 21.957 uTm 14:14:37 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 14:14:41 Integrated Voltage Measurement... 14:14:50 Voltage Offset Correction: v_off_ini = 190.816 uV, v_off_fin = 169.721 uV 14:14:50 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -184.221 uVs 14:14:50 Integrated Voltage Measurement... 14:14:59 Voltage Offset Correction: v_off_ini = 165.762 uV, v_off_fin = 163.130 uV 14:14:59 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 198.143 uVs 14:14:59 Integrated Voltage Measurement... 14:15:08 Voltage Offset Correction: v_off_ini = 148.154 uV, v_off_fin = 124.295 uV 14:15:08 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -192.210 uVs 14:15:08 Integrated Voltage Measurement... 14:15:17 Voltage Offset Correction: v_off_ini = 121.173 uV, v_off_fin = 126.663 uV 14:15:17 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 193.812 uVs 14:15:17 Integrated Voltage Measurement... 14:15:26 Voltage Offset Correction: v_off_ini = -254.351 uV, v_off_fin = 47.465 uV 14:15:26 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 83.783 uVs 14:15:26 Integrated Voltage Measurement... 14:15:35 Voltage Offset Correction: v_off_ini = -473.776 uV, v_off_fin = 13.825 uV 14:15:35 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 604.045 uVs 14:15:39 I1Y Average and RMS of Measurements VT = -214.775 +- 32.080 uVs I1Y = -42.954 +- 6.413 uTm 14:15:45 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 14:15:47 Integrated Voltage Measurement... 14:15:56 Voltage Offset Correction: v_off_ini = 30.973 uV, v_off_fin = 0.770 uV 14:15:56 Integrated Voltage Measurement dXYus = -1.0002 mm, dXYds = 0.0000 mm dXY = -1.0002 mm VT = 6.551 uVs 14:15:56 Integrated Voltage Measurement... 14:16:05 Voltage Offset Correction: v_off_ini = 36.382 uV, v_off_fin = 66.183 uV 14:16:05 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -15.750 uVs 14:16:05 Integrated Voltage Measurement... 14:16:14 Voltage Offset Correction: v_off_ini = 86.985 uV, v_off_fin = 83.640 uV 14:16:14 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -7.865 uVs 14:16:14 Integrated Voltage Measurement... 14:16:23 Voltage Offset Correction: v_off_ini = 109.763 uV, v_off_fin = 93.737 uV 14:16:23 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 14.829 uVs 14:16:23 Integrated Voltage Measurement... 14:16:32 Voltage Offset Correction: v_off_ini = 184.840 uV, v_off_fin = 211.422 uV 14:16:32 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -35.506 uVs 14:16:32 Integrated Voltage Measurement... 14:16:41 Voltage Offset Correction: v_off_ini = 704.349 uV, v_off_fin = 149.711 uV 14:16:41 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -1083.890 uVs 14:16:44 I2X Average and RMS of Measurements VT = 174.665 +- 247.324 uVs I2X = -16.069 +- 22.754 uTm^2 14:16:49 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 14:16:52 Integrated Voltage Measurement... 14:17:01 Voltage Offset Correction: v_off_ini = 131.961 uV, v_off_fin = 126.965 uV 14:17:01 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -427.663 uVs 14:17:01 Integrated Voltage Measurement... 14:17:10 Voltage Offset Correction: v_off_ini = 105.775 uV, v_off_fin = 193.531 uV 14:17:10 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 682.838 uVs 14:17:10 Integrated Voltage Measurement... 14:17:19 Voltage Offset Correction: v_off_ini = 198.791 uV, v_off_fin = 211.369 uV 14:17:19 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -750.505 uVs 14:17:19 Integrated Voltage Measurement... 14:17:28 Voltage Offset Correction: v_off_ini = 244.712 uV, v_off_fin = 225.776 uV 14:17:28 Integrated Voltage Measurement dXYus = -1.0001 mm, dXYds = 0.0001 mm dXY = -1.0001 mm VT = 559.118 uVs 14:17:28 Integrated Voltage Measurement... 14:17:37 Voltage Offset Correction: v_off_ini = 269.451 uV, v_off_fin = 244.514 uV 14:17:37 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -105.145 uVs 14:17:37 Integrated Voltage Measurement... 14:17:46 Voltage Offset Correction: v_off_ini = 264.169 uV, v_off_fin = 195.138 uV 14:17:46 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 511.621 uVs 14:17:48 I2Y Average and RMS of Measurements VT = -506.148 +- 145.628 uVs I2Y = -46.565 +- 13.397 uTm^2