SLAC Magnetic Measurements Date: 11-07-2025 Time: 16:16:15 Project: LCLS-II-HE Magnet Type: Phase Shifter Magnet Name: HE_SXPS_16349 Dataset Number: 1 Run Purpose: Final Results Run Type: Good Field Region Run Number: 1 Measurement Device: Stretched Wire Operator: YL Comment: 16:16:23 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:16:25 Motion Parameters, X Controller, Axis 1 acc = 0.001000 m/s/s dcl = 0.001000 m/s/s vel = 0.001000 m/s pos = 0.000000 m kp = 188.600000 ki = 820.000000 kd = 491.400000 db = 0.000000 16:16:27 Motion Parameters, X Controller, Axis 2 acc = 0.001000 m/s/s dcl = 0.001000 m/s/s vel = 0.001000 m/s pos = 0.000000 m kp = 188.600000 ki = 820.000000 kd = 491.400000 db = 0.000000 16:16:28 Motion Parameters, Y Controller, Axis 1 acc = 0.001000 m/s/s dcl = 0.001000 m/s/s vel = 0.001000 m/s pos = 0.000000 m kp = 188.600000 ki = 820.000000 kd = 491.400000 db = 0.000000 16:16:30 Motion Parameters, Y Controller, Axis 2 acc = 0.001000 m/s/s dcl = 0.001000 m/s/s vel = 0.001000 m/s pos = 0.000000 m kp = 188.600000 ki = 820.000000 kd = 491.400000 db = 0.000000 16:17:08 Wire Position: X0 = -2.0000 mm, Y0 = 0.0000 mm 16:17:12 Integrated Voltage Measurement... 16:17:21 Voltage Offset Correction: v_off_ini = 259.734 uV, v_off_fin = 267.751 uV 16:17:21 Integrated Voltage Measurement dXYus = -0.4998 mm, dXYds = -0.4998 mm dXY = -0.4998 mm VT = -93.816 uVs 16:17:21 Integrated Voltage Measurement... 16:17:30 Voltage Offset Correction: v_off_ini = 275.180 uV, v_off_fin = 266.287 uV 16:17:30 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 89.490 uVs 16:17:30 Integrated Voltage Measurement... 16:17:39 Voltage Offset Correction: v_off_ini = 257.681 uV, v_off_fin = 268.263 uV 16:17:39 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -72.312 uVs 16:17:39 Integrated Voltage Measurement... 16:17:48 Voltage Offset Correction: v_off_ini = 263.556 uV, v_off_fin = 267.476 uV 16:17:48 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 80.998 uVs 16:17:48 Integrated Voltage Measurement... 16:17:57 Voltage Offset Correction: v_off_ini = 261.603 uV, v_off_fin = 266.358 uV 16:17:57 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -90.299 uVs 16:17:57 Integrated Voltage Measurement... 16:18:06 Voltage Offset Correction: v_off_ini = 265.689 uV, v_off_fin = 277.806 uV 16:18:06 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 87.926 uVs 16:18:10 I1X Average and RMS of Measurements VT = -85.807 +- 6.554 uVs I1X = 17.163 +- 1.312 uTm 16:18:13 Wire Position: X0 = -2.0000 mm, Y0 = 0.0000 mm 16:18:17 Integrated Voltage Measurement... 16:18:26 Voltage Offset Correction: v_off_ini = 261.909 uV, v_off_fin = 264.211 uV 16:18:26 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -201.900 uVs 16:18:26 Integrated Voltage Measurement... 16:18:35 Voltage Offset Correction: v_off_ini = 264.313 uV, v_off_fin = 278.774 uV 16:18:35 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 192.333 uVs 16:18:35 Integrated Voltage Measurement... 16:18:44 Voltage Offset Correction: v_off_ini = 273.374 uV, v_off_fin = 266.372 uV 16:18:44 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -196.349 uVs 16:18:44 Integrated Voltage Measurement... 16:18:53 Voltage Offset Correction: v_off_ini = 283.784 uV, v_off_fin = 279.121 uV 16:18:53 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 200.088 uVs 16:18:53 Integrated Voltage Measurement... 16:19:02 Voltage Offset Correction: v_off_ini = 263.628 uV, v_off_fin = 247.815 uV 16:19:02 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -211.203 uVs 16:19:02 Integrated Voltage Measurement... 16:19:11 Voltage Offset Correction: v_off_ini = 273.305 uV, v_off_fin = 286.470 uV 16:19:11 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 194.674 uVs 16:19:15 I1Y Average and RMS of Measurements VT = -199.424 +- 2.525 uVs I1Y = -39.885 +- 0.505 uTm 16:19:20 Wire Position: X0 = -2.0000 mm, Y0 = 0.0000 mm 16:19:23 Integrated Voltage Measurement... 16:19:32 Voltage Offset Correction: v_off_ini = 246.718 uV, v_off_fin = 256.968 uV 16:19:32 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -72.675 uVs 16:19:32 Integrated Voltage Measurement... 16:19:41 Voltage Offset Correction: v_off_ini = 271.816 uV, v_off_fin = 275.202 uV 16:19:41 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 101.144 uVs 16:19:41 Integrated Voltage Measurement... 16:19:50 Voltage Offset Correction: v_off_ini = 262.091 uV, v_off_fin = 263.525 uV 16:19:50 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -79.134 uVs 16:19:50 Integrated Voltage Measurement... 16:19:59 Voltage Offset Correction: v_off_ini = 269.596 uV, v_off_fin = 283.349 uV 16:19:59 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 97.405 uVs 16:19:59 Integrated Voltage Measurement... 16:20:08 Voltage Offset Correction: v_off_ini = 257.374 uV, v_off_fin = 251.387 uV 16:20:08 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -85.249 uVs 16:20:08 Integrated Voltage Measurement... 16:20:17 Voltage Offset Correction: v_off_ini = 268.015 uV, v_off_fin = 263.582 uV 16:20:17 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 102.644 uVs 16:20:19 I2X Average and RMS of Measurements VT = -89.708 +- 3.048 uVs I2X = 8.253 +- 0.280 uTm^2 16:20:24 Wire Position: X0 = -2.0000 mm, Y0 = 0.0000 mm 16:20:27 Integrated Voltage Measurement... 16:20:36 Voltage Offset Correction: v_off_ini = 270.221 uV, v_off_fin = 259.125 uV 16:20:36 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -324.298 uVs 16:20:36 Integrated Voltage Measurement... 16:20:45 Voltage Offset Correction: v_off_ini = 258.518 uV, v_off_fin = 252.773 uV 16:20:45 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 328.652 uVs 16:20:45 Integrated Voltage Measurement... 16:20:54 Voltage Offset Correction: v_off_ini = 247.443 uV, v_off_fin = 256.341 uV 16:20:54 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -318.180 uVs 16:20:54 Integrated Voltage Measurement... 16:21:03 Voltage Offset Correction: v_off_ini = 258.845 uV, v_off_fin = 260.160 uV 16:21:03 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 320.116 uVs 16:21:03 Integrated Voltage Measurement... 16:21:12 Voltage Offset Correction: v_off_ini = 252.840 uV, v_off_fin = 259.609 uV 16:21:12 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -325.718 uVs 16:21:12 Integrated Voltage Measurement... 16:21:21 Voltage Offset Correction: v_off_ini = 260.121 uV, v_off_fin = 265.203 uV 16:21:21 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 331.170 uVs 16:21:23 I2Y Average and RMS of Measurements VT = -324.689 +- 3.999 uVs I2Y = -29.871 +- 0.368 uTm^2 16:21:28 Wire Position: X0 = -1.0000 mm, Y0 = 0.0000 mm 16:21:32 Integrated Voltage Measurement... 16:21:41 Voltage Offset Correction: v_off_ini = 257.779 uV, v_off_fin = 254.154 uV 16:21:41 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -64.333 uVs 16:21:41 Integrated Voltage Measurement... 16:21:50 Voltage Offset Correction: v_off_ini = 248.060 uV, v_off_fin = 257.771 uV 16:21:50 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 58.932 uVs 16:21:50 Integrated Voltage Measurement... 16:21:59 Voltage Offset Correction: v_off_ini = 262.812 uV, v_off_fin = 250.677 uV 16:21:59 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -62.422 uVs 16:21:59 Integrated Voltage Measurement... 16:22:08 Voltage Offset Correction: v_off_ini = 267.402 uV, v_off_fin = 264.005 uV 16:22:08 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 46.023 uVs 16:22:08 Integrated Voltage Measurement... 16:22:17 Voltage Offset Correction: v_off_ini = 260.388 uV, v_off_fin = 258.634 uV 16:22:17 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -67.738 uVs 16:22:17 Integrated Voltage Measurement... 16:22:26 Voltage Offset Correction: v_off_ini = 262.604 uV, v_off_fin = 265.329 uV 16:22:26 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 71.789 uVs 16:22:30 I1X Average and RMS of Measurements VT = -61.873 +- 6.347 uVs I1X = 12.375 +- 1.269 uTm 16:22:33 Wire Position: X0 = -1.0000 mm, Y0 = 0.0000 mm 16:22:37 Integrated Voltage Measurement... 16:22:46 Voltage Offset Correction: v_off_ini = 261.601 uV, v_off_fin = 251.560 uV 16:22:46 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -198.812 uVs 16:22:46 Integrated Voltage Measurement... 16:22:55 Voltage Offset Correction: v_off_ini = 265.930 uV, v_off_fin = 272.025 uV 16:22:55 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 179.829 uVs 16:22:55 Integrated Voltage Measurement... 16:23:04 Voltage Offset Correction: v_off_ini = 259.387 uV, v_off_fin = 238.374 uV 16:23:04 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -220.989 uVs 16:23:04 Integrated Voltage Measurement... 16:23:13 Voltage Offset Correction: v_off_ini = 239.048 uV, v_off_fin = 248.988 uV 16:23:13 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 222.966 uVs 16:23:13 Integrated Voltage Measurement... 16:23:22 Voltage Offset Correction: v_off_ini = 247.100 uV, v_off_fin = 224.321 uV 16:23:22 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -223.835 uVs 16:23:22 Integrated Voltage Measurement... 16:23:31 Voltage Offset Correction: v_off_ini = 241.233 uV, v_off_fin = 254.359 uV 16:23:31 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 229.546 uVs 16:23:35 I1Y Average and RMS of Measurements VT = -212.663 +- 16.617 uVs I1Y = -42.533 +- 3.323 uTm 16:23:40 Wire Position: X0 = -1.0000 mm, Y0 = 0.0000 mm 16:23:43 Integrated Voltage Measurement... 16:23:52 Voltage Offset Correction: v_off_ini = 237.097 uV, v_off_fin = 227.744 uV 16:23:52 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -64.631 uVs 16:23:52 Integrated Voltage Measurement... 16:24:01 Voltage Offset Correction: v_off_ini = 231.832 uV, v_off_fin = 249.134 uV 16:24:01 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 91.766 uVs 16:24:01 Integrated Voltage Measurement... 16:24:10 Voltage Offset Correction: v_off_ini = 238.831 uV, v_off_fin = 230.552 uV 16:24:10 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -76.277 uVs 16:24:10 Integrated Voltage Measurement... 16:24:19 Voltage Offset Correction: v_off_ini = 229.988 uV, v_off_fin = 247.234 uV 16:24:19 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 83.961 uVs 16:24:19 Integrated Voltage Measurement... 16:24:28 Voltage Offset Correction: v_off_ini = 243.896 uV, v_off_fin = 222.063 uV 16:24:28 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -85.825 uVs 16:24:28 Integrated Voltage Measurement... 16:24:37 Voltage Offset Correction: v_off_ini = 241.969 uV, v_off_fin = 240.778 uV 16:24:37 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 71.604 uVs 16:24:39 I2X Average and RMS of Measurements VT = -79.011 +- 0.812 uVs I2X = 7.269 +- 0.075 uTm^2 16:24:44 Wire Position: X0 = -1.0000 mm, Y0 = 0.0000 mm 16:24:47 Integrated Voltage Measurement... 16:24:56 Voltage Offset Correction: v_off_ini = 254.429 uV, v_off_fin = 242.836 uV 16:24:56 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -324.500 uVs 16:24:56 Integrated Voltage Measurement... 16:25:05 Voltage Offset Correction: v_off_ini = 246.033 uV, v_off_fin = 239.341 uV 16:25:05 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 341.986 uVs 16:25:05 Integrated Voltage Measurement... 16:25:14 Voltage Offset Correction: v_off_ini = 250.142 uV, v_off_fin = 243.087 uV 16:25:14 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -337.155 uVs 16:25:14 Integrated Voltage Measurement... 16:25:23 Voltage Offset Correction: v_off_ini = 243.196 uV, v_off_fin = 239.106 uV 16:25:23 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 342.093 uVs 16:25:23 Integrated Voltage Measurement... 16:25:32 Voltage Offset Correction: v_off_ini = 231.912 uV, v_off_fin = 244.749 uV 16:25:32 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -338.942 uVs 16:25:32 Integrated Voltage Measurement... 16:25:41 Voltage Offset Correction: v_off_ini = 241.448 uV, v_off_fin = 241.720 uV 16:25:41 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 340.441 uVs 16:25:43 I2Y Average and RMS of Measurements VT = -337.520 +- 3.024 uVs I2Y = -31.052 +- 0.278 uTm^2 16:25:48 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:25:52 Integrated Voltage Measurement... 16:26:01 Voltage Offset Correction: v_off_ini = 249.681 uV, v_off_fin = 246.233 uV 16:26:01 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.4998 mm dXY = -0.4999 mm VT = -69.871 uVs 16:26:01 Integrated Voltage Measurement... 16:26:10 Voltage Offset Correction: v_off_ini = 236.527 uV, v_off_fin = 247.870 uV 16:26:10 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 15.241 uVs 16:26:10 Integrated Voltage Measurement... 16:26:19 Voltage Offset Correction: v_off_ini = 247.059 uV, v_off_fin = 234.121 uV 16:26:19 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -61.107 uVs 16:26:19 Integrated Voltage Measurement... 16:26:28 Voltage Offset Correction: v_off_ini = 254.065 uV, v_off_fin = 250.390 uV 16:26:28 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 39.523 uVs 16:26:28 Integrated Voltage Measurement... 16:26:37 Voltage Offset Correction: v_off_ini = 226.874 uV, v_off_fin = 227.990 uV 16:26:37 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -42.125 uVs 16:26:37 Integrated Voltage Measurement... 16:26:46 Voltage Offset Correction: v_off_ini = 239.716 uV, v_off_fin = 224.061 uV 16:26:46 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 52.315 uVs 16:26:50 I1X Average and RMS of Measurements VT = -46.697 +- 3.189 uVs I1X = 9.340 +- 0.637 uTm 16:26:53 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:26:57 Integrated Voltage Measurement... 16:27:06 Voltage Offset Correction: v_off_ini = 219.614 uV, v_off_fin = 215.573 uV 16:27:06 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -210.152 uVs 16:27:06 Integrated Voltage Measurement... 16:27:15 Voltage Offset Correction: v_off_ini = 224.385 uV, v_off_fin = 227.141 uV 16:27:15 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 212.620 uVs 16:27:15 Integrated Voltage Measurement... 16:27:24 Voltage Offset Correction: v_off_ini = 219.854 uV, v_off_fin = 220.623 uV 16:27:24 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -208.176 uVs 16:27:24 Integrated Voltage Measurement... 16:27:33 Voltage Offset Correction: v_off_ini = 228.282 uV, v_off_fin = 232.012 uV 16:27:33 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 213.057 uVs 16:27:33 Integrated Voltage Measurement... 16:27:42 Voltage Offset Correction: v_off_ini = 237.970 uV, v_off_fin = 226.964 uV 16:27:42 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -204.653 uVs 16:27:42 Integrated Voltage Measurement... 16:27:51 Voltage Offset Correction: v_off_ini = 242.247 uV, v_off_fin = 256.143 uV 16:27:51 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 191.137 uVs 16:27:55 I1Y Average and RMS of Measurements VT = -206.632 +- 6.186 uVs I1Y = -41.326 +- 1.237 uTm 16:28:00 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:28:03 Integrated Voltage Measurement... 16:28:12 Voltage Offset Correction: v_off_ini = 223.249 uV, v_off_fin = 240.110 uV 16:28:12 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -48.203 uVs 16:28:12 Integrated Voltage Measurement... 16:28:21 Voltage Offset Correction: v_off_ini = 241.975 uV, v_off_fin = 245.642 uV 16:28:21 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 75.553 uVs 16:28:21 Integrated Voltage Measurement... 16:28:30 Voltage Offset Correction: v_off_ini = 232.845 uV, v_off_fin = 226.070 uV 16:28:30 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -47.279 uVs 16:28:30 Integrated Voltage Measurement... 16:28:39 Voltage Offset Correction: v_off_ini = 233.152 uV, v_off_fin = 240.678 uV 16:28:39 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 70.557 uVs 16:28:39 Integrated Voltage Measurement... 16:28:48 Voltage Offset Correction: v_off_ini = 219.474 uV, v_off_fin = 225.414 uV 16:28:48 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -58.085 uVs 16:28:48 Integrated Voltage Measurement... 16:28:57 Voltage Offset Correction: v_off_ini = 245.315 uV, v_off_fin = 257.815 uV 16:28:57 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 64.958 uVs 16:29:00 I2X Average and RMS of Measurements VT = -60.772 +- 1.320 uVs I2X = 5.591 +- 0.121 uTm^2 16:29:05 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:29:07 Integrated Voltage Measurement... 16:29:16 Voltage Offset Correction: v_off_ini = 234.701 uV, v_off_fin = 239.185 uV 16:29:16 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -337.177 uVs 16:29:16 Integrated Voltage Measurement... 16:29:25 Voltage Offset Correction: v_off_ini = 251.380 uV, v_off_fin = 239.906 uV 16:29:25 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 331.063 uVs 16:29:25 Integrated Voltage Measurement... 16:29:34 Voltage Offset Correction: v_off_ini = 232.224 uV, v_off_fin = 238.447 uV 16:29:34 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -327.527 uVs 16:29:34 Integrated Voltage Measurement... 16:29:43 Voltage Offset Correction: v_off_ini = 232.544 uV, v_off_fin = 225.468 uV 16:29:43 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 332.547 uVs 16:29:43 Integrated Voltage Measurement... 16:29:52 Voltage Offset Correction: v_off_ini = 236.433 uV, v_off_fin = 229.752 uV 16:29:52 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -322.821 uVs 16:29:52 Integrated Voltage Measurement... 16:30:01 Voltage Offset Correction: v_off_ini = 234.177 uV, v_off_fin = 235.416 uV 16:30:01 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 326.860 uVs 16:30:04 I2Y Average and RMS of Measurements VT = -329.666 +- 3.797 uVs I2Y = -30.329 +- 0.349 uTm^2 16:30:08 Wire Position: X0 = 1.0000 mm, Y0 = 0.0000 mm 16:30:13 Integrated Voltage Measurement... 16:30:22 Voltage Offset Correction: v_off_ini = 223.937 uV, v_off_fin = 227.878 uV 16:30:22 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.4999 mm dXY = -0.4999 mm VT = -45.310 uVs 16:30:22 Integrated Voltage Measurement... 16:30:31 Voltage Offset Correction: v_off_ini = 221.381 uV, v_off_fin = 241.775 uV 16:30:31 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 48.919 uVs 16:30:31 Integrated Voltage Measurement... 16:30:40 Voltage Offset Correction: v_off_ini = 240.884 uV, v_off_fin = 238.524 uV 16:30:40 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -41.351 uVs 16:30:40 Integrated Voltage Measurement... 16:30:49 Voltage Offset Correction: v_off_ini = 234.922 uV, v_off_fin = 239.039 uV 16:30:49 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 38.388 uVs 16:30:49 Integrated Voltage Measurement... 16:30:58 Voltage Offset Correction: v_off_ini = 231.433 uV, v_off_fin = 234.610 uV 16:30:58 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -63.701 uVs 16:30:58 Integrated Voltage Measurement... 16:31:07 Voltage Offset Correction: v_off_ini = 242.635 uV, v_off_fin = 245.112 uV 16:31:07 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 40.781 uVs 16:31:11 I1X Average and RMS of Measurements VT = -46.408 +- 5.075 uVs I1X = 9.282 +- 1.015 uTm 16:31:13 Wire Position: X0 = 1.0000 mm, Y0 = 0.0000 mm 16:31:18 Integrated Voltage Measurement... 16:31:26 Voltage Offset Correction: v_off_ini = 240.412 uV, v_off_fin = 230.237 uV 16:31:26 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -212.027 uVs 16:31:27 Integrated Voltage Measurement... 16:31:35 Voltage Offset Correction: v_off_ini = 244.618 uV, v_off_fin = 253.416 uV 16:31:35 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 196.327 uVs 16:31:36 Integrated Voltage Measurement... 16:31:44 Voltage Offset Correction: v_off_ini = 237.715 uV, v_off_fin = 245.386 uV 16:31:44 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -210.514 uVs 16:31:44 Integrated Voltage Measurement... 16:31:53 Voltage Offset Correction: v_off_ini = 246.175 uV, v_off_fin = 258.087 uV 16:31:53 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 200.910 uVs 16:31:53 Integrated Voltage Measurement... 16:32:02 Voltage Offset Correction: v_off_ini = 243.493 uV, v_off_fin = 232.435 uV 16:32:02 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -205.539 uVs 16:32:02 Integrated Voltage Measurement... 16:32:11 Voltage Offset Correction: v_off_ini = 240.787 uV, v_off_fin = 255.820 uV 16:32:11 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 211.421 uVs 16:32:16 I1Y Average and RMS of Measurements VT = -206.123 +- 1.780 uVs I1Y = -41.225 +- 0.356 uTm 16:32:21 Wire Position: X0 = 1.0000 mm, Y0 = 0.0000 mm 16:32:23 Integrated Voltage Measurement... 16:32:32 Voltage Offset Correction: v_off_ini = 248.558 uV, v_off_fin = 236.427 uV 16:32:32 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -43.501 uVs 16:32:32 Integrated Voltage Measurement... 16:32:41 Voltage Offset Correction: v_off_ini = 238.586 uV, v_off_fin = 253.430 uV 16:32:41 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 69.382 uVs 16:32:41 Integrated Voltage Measurement... 16:32:50 Voltage Offset Correction: v_off_ini = 248.100 uV, v_off_fin = 242.294 uV 16:32:50 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -45.446 uVs 16:32:50 Integrated Voltage Measurement... 16:32:59 Voltage Offset Correction: v_off_ini = 241.508 uV, v_off_fin = 247.325 uV 16:32:59 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 56.191 uVs 16:32:59 Integrated Voltage Measurement... 16:33:08 Voltage Offset Correction: v_off_ini = 243.445 uV, v_off_fin = 240.300 uV 16:33:08 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -38.403 uVs 16:33:08 Integrated Voltage Measurement... 16:33:17 Voltage Offset Correction: v_off_ini = 247.968 uV, v_off_fin = 260.063 uV 16:33:17 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 39.228 uVs 16:33:20 I2X Average and RMS of Measurements VT = -48.692 +- 7.351 uVs I2X = 4.480 +- 0.676 uTm^2 16:33:25 Wire Position: X0 = 1.0000 mm, Y0 = 0.0000 mm 16:33:28 Integrated Voltage Measurement... 16:33:37 Voltage Offset Correction: v_off_ini = 253.978 uV, v_off_fin = 248.923 uV 16:33:37 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -346.166 uVs 16:33:37 Integrated Voltage Measurement... 16:33:46 Voltage Offset Correction: v_off_ini = 250.685 uV, v_off_fin = 260.642 uV 16:33:46 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 313.519 uVs 16:33:46 Integrated Voltage Measurement... 16:33:55 Voltage Offset Correction: v_off_ini = 244.931 uV, v_off_fin = 250.398 uV 16:33:55 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -346.716 uVs 16:33:55 Integrated Voltage Measurement... 16:34:04 Voltage Offset Correction: v_off_ini = 258.199 uV, v_off_fin = 260.015 uV 16:34:04 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 336.633 uVs 16:34:04 Integrated Voltage Measurement... 16:34:13 Voltage Offset Correction: v_off_ini = 250.863 uV, v_off_fin = 255.057 uV 16:34:13 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -327.018 uVs 16:34:13 Integrated Voltage Measurement... 16:34:22 Voltage Offset Correction: v_off_ini = 249.947 uV, v_off_fin = 256.537 uV 16:34:22 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 331.963 uVs 16:34:24 I2Y Average and RMS of Measurements VT = -333.669 +- 5.662 uVs I2Y = -30.698 +- 0.521 uTm^2 16:34:29 Wire Position: X0 = 2.0000 mm, Y0 = 0.0000 mm 16:34:33 Integrated Voltage Measurement... 16:34:42 Voltage Offset Correction: v_off_ini = 259.075 uV, v_off_fin = 251.870 uV 16:34:42 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.4998 mm dXY = -0.4999 mm VT = -40.436 uVs 16:34:42 Integrated Voltage Measurement... 16:34:51 Voltage Offset Correction: v_off_ini = 242.297 uV, v_off_fin = 236.566 uV 16:34:51 Integrated Voltage Measurement dXYus = 0.4998 mm, dXYds = 0.5000 mm dXY = 0.4999 mm VT = 34.493 uVs 16:34:51 Integrated Voltage Measurement... 16:35:00 Voltage Offset Correction: v_off_ini = 242.207 uV, v_off_fin = 237.090 uV 16:35:00 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -39.504 uVs 16:35:00 Integrated Voltage Measurement... 16:35:09 Voltage Offset Correction: v_off_ini = 225.837 uV, v_off_fin = 249.804 uV 16:35:09 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 33.134 uVs 16:35:09 Integrated Voltage Measurement... 16:35:18 Voltage Offset Correction: v_off_ini = 249.779 uV, v_off_fin = 249.739 uV 16:35:18 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -49.263 uVs 16:35:18 Integrated Voltage Measurement... 16:35:27 Voltage Offset Correction: v_off_ini = 238.594 uV, v_off_fin = 237.718 uV 16:35:27 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 52.144 uVs 16:35:31 I1X Average and RMS of Measurements VT = -41.496 +- 6.528 uVs I1X = 8.300 +- 1.305 uTm 16:35:34 Wire Position: X0 = 2.0000 mm, Y0 = 0.0000 mm 16:35:38 Integrated Voltage Measurement... 16:35:47 Voltage Offset Correction: v_off_ini = 252.798 uV, v_off_fin = 233.448 uV 16:35:47 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -204.882 uVs 16:35:47 Integrated Voltage Measurement... 16:35:56 Voltage Offset Correction: v_off_ini = 235.525 uV, v_off_fin = 244.816 uV 16:35:56 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 202.647 uVs 16:35:56 Integrated Voltage Measurement... 16:36:05 Voltage Offset Correction: v_off_ini = 235.023 uV, v_off_fin = 219.101 uV 16:36:05 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -191.685 uVs 16:36:05 Integrated Voltage Measurement... 16:36:14 Voltage Offset Correction: v_off_ini = 227.240 uV, v_off_fin = 238.467 uV 16:36:14 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 192.657 uVs 16:36:14 Integrated Voltage Measurement... 16:36:23 Voltage Offset Correction: v_off_ini = 230.642 uV, v_off_fin = 209.650 uV 16:36:23 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -217.441 uVs 16:36:23 Integrated Voltage Measurement... 16:36:32 Voltage Offset Correction: v_off_ini = 233.706 uV, v_off_fin = 226.263 uV 16:36:32 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 189.210 uVs 16:36:36 I1Y Average and RMS of Measurements VT = -199.753 +- 5.365 uVs I1Y = -39.951 +- 1.073 uTm 16:36:41 Wire Position: X0 = 2.0000 mm, Y0 = 0.0000 mm 16:36:44 Integrated Voltage Measurement... 16:36:53 Voltage Offset Correction: v_off_ini = 218.033 uV, v_off_fin = 205.086 uV 16:36:53 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -42.682 uVs 16:36:53 Integrated Voltage Measurement... 16:37:02 Voltage Offset Correction: v_off_ini = 212.591 uV, v_off_fin = 226.281 uV 16:37:02 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 47.732 uVs 16:37:02 Integrated Voltage Measurement... 16:37:11 Voltage Offset Correction: v_off_ini = 226.590 uV, v_off_fin = 209.879 uV 16:37:11 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -54.333 uVs 16:37:11 Integrated Voltage Measurement... 16:37:20 Voltage Offset Correction: v_off_ini = 228.070 uV, v_off_fin = 232.331 uV 16:37:20 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 49.549 uVs 16:37:20 Integrated Voltage Measurement... 16:37:29 Voltage Offset Correction: v_off_ini = 219.385 uV, v_off_fin = 211.676 uV 16:37:29 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -39.195 uVs 16:37:29 Integrated Voltage Measurement... 16:37:38 Voltage Offset Correction: v_off_ini = 210.163 uV, v_off_fin = 213.357 uV 16:37:38 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 54.303 uVs 16:37:40 I2X Average and RMS of Measurements VT = -47.966 +- 2.881 uVs I2X = 4.413 +- 0.265 uTm^2 16:37:45 Wire Position: X0 = 2.0000 mm, Y0 = 0.0000 mm 16:37:48 Integrated Voltage Measurement... 16:37:57 Voltage Offset Correction: v_off_ini = 204.938 uV, v_off_fin = 207.016 uV 16:37:57 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -324.124 uVs 16:37:57 Integrated Voltage Measurement... 16:38:06 Voltage Offset Correction: v_off_ini = 208.102 uV, v_off_fin = 216.655 uV 16:38:06 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 323.704 uVs 16:38:06 Integrated Voltage Measurement... 16:38:15 Voltage Offset Correction: v_off_ini = 201.030 uV, v_off_fin = 207.285 uV 16:38:15 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -325.394 uVs 16:38:15 Integrated Voltage Measurement... 16:38:24 Voltage Offset Correction: v_off_ini = 201.345 uV, v_off_fin = 215.513 uV 16:38:24 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 307.984 uVs 16:38:24 Integrated Voltage Measurement... 16:38:33 Voltage Offset Correction: v_off_ini = 203.973 uV, v_off_fin = 199.709 uV 16:38:33 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -299.324 uVs 16:38:33 Integrated Voltage Measurement... 16:38:42 Voltage Offset Correction: v_off_ini = 187.836 uV, v_off_fin = 200.796 uV 16:38:42 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 325.717 uVs 16:38:45 I2Y Average and RMS of Measurements VT = -317.708 +- 4.707 uVs I2Y = -29.229 +- 0.433 uTm^2 16:38:52 Wire Position: X0 = 0.0000 mm, Y0 = -1.0000 mm 16:38:56 Integrated Voltage Measurement... 16:39:05 Voltage Offset Correction: v_off_ini = 224.327 uV, v_off_fin = 212.937 uV 16:39:05 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -48.763 uVs 16:39:05 Integrated Voltage Measurement... 16:39:14 Voltage Offset Correction: v_off_ini = 212.689 uV, v_off_fin = 215.833 uV 16:39:14 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 53.456 uVs 16:39:14 Integrated Voltage Measurement... 16:39:23 Voltage Offset Correction: v_off_ini = 218.094 uV, v_off_fin = 215.578 uV 16:39:23 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -37.010 uVs 16:39:23 Integrated Voltage Measurement... 16:39:32 Voltage Offset Correction: v_off_ini = 226.789 uV, v_off_fin = 218.150 uV 16:39:32 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 61.021 uVs 16:39:33 Integrated Voltage Measurement... 16:39:41 Voltage Offset Correction: v_off_ini = 211.604 uV, v_off_fin = 208.731 uV 16:39:41 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -31.546 uVs 16:39:42 Integrated Voltage Measurement... 16:39:50 Voltage Offset Correction: v_off_ini = 214.119 uV, v_off_fin = 224.805 uV 16:39:50 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 37.972 uVs 16:39:55 I1X Average and RMS of Measurements VT = -44.961 +- 7.265 uVs I1X = 8.992 +- 1.453 uTm 16:39:57 Wire Position: X0 = 0.0000 mm, Y0 = -1.0000 mm 16:40:01 Integrated Voltage Measurement... 16:40:10 Voltage Offset Correction: v_off_ini = 214.950 uV, v_off_fin = 201.260 uV 16:40:10 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -212.273 uVs 16:40:10 Integrated Voltage Measurement... 16:40:19 Voltage Offset Correction: v_off_ini = 197.961 uV, v_off_fin = 227.791 uV 16:40:19 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 224.313 uVs 16:40:19 Integrated Voltage Measurement... 16:40:28 Voltage Offset Correction: v_off_ini = 199.393 uV, v_off_fin = 201.108 uV 16:40:28 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -224.797 uVs 16:40:28 Integrated Voltage Measurement... 16:40:37 Voltage Offset Correction: v_off_ini = 206.589 uV, v_off_fin = 227.662 uV 16:40:37 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 222.069 uVs 16:40:37 Integrated Voltage Measurement... 16:40:46 Voltage Offset Correction: v_off_ini = 217.124 uV, v_off_fin = 203.025 uV 16:40:46 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -224.864 uVs 16:40:46 Integrated Voltage Measurement... 16:40:55 Voltage Offset Correction: v_off_ini = 206.853 uV, v_off_fin = 224.679 uV 16:40:55 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 224.492 uVs 16:40:59 I1Y Average and RMS of Measurements VT = -222.135 +- 2.764 uVs I1Y = -44.427 +- 0.553 uTm 16:41:02 Wire Position: X0 = 0.0000 mm, Y0 = -1.0000 mm 16:41:05 Integrated Voltage Measurement... 16:41:14 Voltage Offset Correction: v_off_ini = 204.799 uV, v_off_fin = 207.429 uV 16:41:14 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -43.199 uVs 16:41:14 Integrated Voltage Measurement... 16:41:23 Voltage Offset Correction: v_off_ini = 207.329 uV, v_off_fin = 230.374 uV 16:41:23 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 63.204 uVs 16:41:23 Integrated Voltage Measurement... 16:41:32 Voltage Offset Correction: v_off_ini = 219.032 uV, v_off_fin = 212.291 uV 16:41:32 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -32.228 uVs 16:41:32 Integrated Voltage Measurement... 16:41:41 Voltage Offset Correction: v_off_ini = 227.834 uV, v_off_fin = 240.396 uV 16:41:41 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 44.891 uVs 16:41:41 Integrated Voltage Measurement... 16:41:50 Voltage Offset Correction: v_off_ini = 214.373 uV, v_off_fin = 217.013 uV 16:41:50 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -51.218 uVs 16:41:50 Integrated Voltage Measurement... 16:41:59 Voltage Offset Correction: v_off_ini = 219.767 uV, v_off_fin = 223.154 uV 16:41:59 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 53.210 uVs 16:42:01 I2X Average and RMS of Measurements VT = -47.992 +- 6.682 uVs I2X = 4.415 +- 0.615 uTm^2 16:42:03 Wire Position: X0 = 0.0000 mm, Y0 = -1.0000 mm 16:42:06 Integrated Voltage Measurement... 16:42:15 Voltage Offset Correction: v_off_ini = 231.527 uV, v_off_fin = 222.201 uV 16:42:15 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -360.116 uVs 16:42:15 Integrated Voltage Measurement... 16:42:24 Voltage Offset Correction: v_off_ini = 222.302 uV, v_off_fin = 227.033 uV 16:42:24 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 352.989 uVs 16:42:24 Integrated Voltage Measurement... 16:42:33 Voltage Offset Correction: v_off_ini = 219.363 uV, v_off_fin = 218.779 uV 16:42:33 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = -350.623 uVs 16:42:33 Integrated Voltage Measurement... 16:42:42 Voltage Offset Correction: v_off_ini = 218.699 uV, v_off_fin = 220.861 uV 16:42:42 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 360.720 uVs 16:42:42 Integrated Voltage Measurement... 16:42:51 Voltage Offset Correction: v_off_ini = 221.939 uV, v_off_fin = 219.669 uV 16:42:51 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -341.139 uVs 16:42:51 Integrated Voltage Measurement... 16:43:00 Voltage Offset Correction: v_off_ini = 231.106 uV, v_off_fin = 234.980 uV 16:43:00 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 355.459 uVs 16:43:02 I2Y Average and RMS of Measurements VT = -353.508 +- 3.700 uVs I2Y = -32.523 +- 0.341 uTm^2 16:43:06 Wire Position: X0 = 0.0000 mm, Y0 = -0.5000 mm 16:43:11 Integrated Voltage Measurement... 16:43:20 Voltage Offset Correction: v_off_ini = 239.931 uV, v_off_fin = 238.823 uV 16:43:20 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -36.757 uVs 16:43:20 Integrated Voltage Measurement... 16:43:29 Voltage Offset Correction: v_off_ini = 247.412 uV, v_off_fin = 241.787 uV 16:43:29 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 15.121 uVs 16:43:29 Integrated Voltage Measurement... 16:43:38 Voltage Offset Correction: v_off_ini = 254.572 uV, v_off_fin = 251.443 uV 16:43:38 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -50.465 uVs 16:43:38 Integrated Voltage Measurement... 16:43:47 Voltage Offset Correction: v_off_ini = 253.343 uV, v_off_fin = 256.723 uV 16:43:47 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 48.027 uVs 16:43:47 Integrated Voltage Measurement... 16:43:56 Voltage Offset Correction: v_off_ini = 255.950 uV, v_off_fin = 248.665 uV 16:43:56 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.4999 mm dXY = -0.4999 mm VT = -58.506 uVs 16:43:56 Integrated Voltage Measurement... 16:44:05 Voltage Offset Correction: v_off_ini = 240.960 uV, v_off_fin = 244.634 uV 16:44:05 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = 61.015 uVs 16:44:09 I1X Average and RMS of Measurements VT = -44.982 +- 14.133 uVs I1X = 8.997 +- 2.827 uTm 16:44:11 Wire Position: X0 = 0.0000 mm, Y0 = -0.5000 mm 16:44:16 Integrated Voltage Measurement... 16:44:24 Voltage Offset Correction: v_off_ini = 229.122 uV, v_off_fin = 220.716 uV 16:44:25 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -212.111 uVs 16:44:25 Integrated Voltage Measurement... 16:44:33 Voltage Offset Correction: v_off_ini = 226.327 uV, v_off_fin = 242.906 uV 16:44:33 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 210.379 uVs 16:44:34 Integrated Voltage Measurement... 16:44:42 Voltage Offset Correction: v_off_ini = 227.831 uV, v_off_fin = 225.652 uV 16:44:42 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -210.728 uVs 16:44:42 Integrated Voltage Measurement... 16:44:51 Voltage Offset Correction: v_off_ini = 224.567 uV, v_off_fin = 248.657 uV 16:44:51 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 191.298 uVs 16:44:51 Integrated Voltage Measurement... 16:45:00 Voltage Offset Correction: v_off_ini = 232.709 uV, v_off_fin = 220.599 uV 16:45:00 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -207.171 uVs 16:45:00 Integrated Voltage Measurement... 16:45:09 Voltage Offset Correction: v_off_ini = 213.378 uV, v_off_fin = 231.741 uV 16:45:09 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 202.088 uVs 16:45:14 I1Y Average and RMS of Measurements VT = -205.629 +- 4.237 uVs I1Y = -41.126 +- 0.847 uTm 16:45:16 Wire Position: X0 = 0.0000 mm, Y0 = -0.5000 mm 16:45:19 Integrated Voltage Measurement... 16:45:28 Voltage Offset Correction: v_off_ini = 234.664 uV, v_off_fin = 219.911 uV 16:45:28 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -55.893 uVs 16:45:28 Integrated Voltage Measurement... 16:45:37 Voltage Offset Correction: v_off_ini = 226.932 uV, v_off_fin = 245.989 uV 16:45:37 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 55.965 uVs 16:45:37 Integrated Voltage Measurement... 16:45:46 Voltage Offset Correction: v_off_ini = 237.730 uV, v_off_fin = 227.450 uV 16:45:46 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -72.056 uVs 16:45:46 Integrated Voltage Measurement... 16:45:55 Voltage Offset Correction: v_off_ini = 220.378 uV, v_off_fin = 241.580 uV 16:45:55 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 58.221 uVs 16:45:55 Integrated Voltage Measurement... 16:46:04 Voltage Offset Correction: v_off_ini = 228.290 uV, v_off_fin = 221.387 uV 16:46:04 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -67.337 uVs 16:46:04 Integrated Voltage Measurement... 16:46:13 Voltage Offset Correction: v_off_ini = 226.605 uV, v_off_fin = 246.802 uV 16:46:13 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 60.080 uVs 16:46:15 I2X Average and RMS of Measurements VT = -61.592 +- 4.047 uVs I2X = 5.667 +- 0.372 uTm^2 16:46:17 Wire Position: X0 = 0.0000 mm, Y0 = -0.5000 mm 16:46:20 Integrated Voltage Measurement... 16:46:29 Voltage Offset Correction: v_off_ini = 231.549 uV, v_off_fin = 225.970 uV 16:46:29 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -336.863 uVs 16:46:29 Integrated Voltage Measurement... 16:46:38 Voltage Offset Correction: v_off_ini = 243.957 uV, v_off_fin = 237.754 uV 16:46:38 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 335.657 uVs 16:46:38 Integrated Voltage Measurement... 16:46:47 Voltage Offset Correction: v_off_ini = 241.700 uV, v_off_fin = 240.856 uV 16:46:47 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -338.790 uVs 16:46:47 Integrated Voltage Measurement... 16:46:56 Voltage Offset Correction: v_off_ini = 233.137 uV, v_off_fin = 240.232 uV 16:46:56 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 354.252 uVs 16:46:56 Integrated Voltage Measurement... 16:47:05 Voltage Offset Correction: v_off_ini = 237.123 uV, v_off_fin = 219.816 uV 16:47:05 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -325.930 uVs 16:47:05 Integrated Voltage Measurement... 16:47:14 Voltage Offset Correction: v_off_ini = 229.726 uV, v_off_fin = 219.977 uV 16:47:14 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 324.487 uVs 16:47:17 I2Y Average and RMS of Measurements VT = -335.997 +- 8.703 uVs I2Y = -30.912 +- 0.801 uTm^2 16:47:21 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:47:25 Integrated Voltage Measurement... 16:47:34 Voltage Offset Correction: v_off_ini = 234.314 uV, v_off_fin = 235.031 uV 16:47:34 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -62.980 uVs 16:47:34 Integrated Voltage Measurement... 16:47:43 Voltage Offset Correction: v_off_ini = 239.231 uV, v_off_fin = 232.750 uV 16:47:43 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 47.341 uVs 16:47:43 Integrated Voltage Measurement... 16:47:52 Voltage Offset Correction: v_off_ini = 232.684 uV, v_off_fin = 219.328 uV 16:47:52 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -60.207 uVs 16:47:52 Integrated Voltage Measurement... 16:48:01 Voltage Offset Correction: v_off_ini = 218.842 uV, v_off_fin = 216.471 uV 16:48:01 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 58.107 uVs 16:48:01 Integrated Voltage Measurement... 16:48:10 Voltage Offset Correction: v_off_ini = 220.317 uV, v_off_fin = 230.034 uV 16:48:10 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -65.953 uVs 16:48:10 Integrated Voltage Measurement... 16:48:19 Voltage Offset Correction: v_off_ini = 235.704 uV, v_off_fin = 238.121 uV 16:48:19 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 50.528 uVs 16:48:23 I1X Average and RMS of Measurements VT = -57.519 +- 1.709 uVs I1X = 11.504 +- 0.342 uTm 16:48:25 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:48:30 Integrated Voltage Measurement... 16:48:39 Voltage Offset Correction: v_off_ini = 222.970 uV, v_off_fin = 200.901 uV 16:48:39 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -202.876 uVs 16:48:39 Integrated Voltage Measurement... 16:48:48 Voltage Offset Correction: v_off_ini = 196.856 uV, v_off_fin = 211.435 uV 16:48:48 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 214.253 uVs 16:48:48 Integrated Voltage Measurement... 16:48:56 Voltage Offset Correction: v_off_ini = 208.637 uV, v_off_fin = 206.130 uV 16:48:57 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -210.903 uVs 16:48:57 Integrated Voltage Measurement... 16:49:05 Voltage Offset Correction: v_off_ini = 228.177 uV, v_off_fin = 230.648 uV 16:49:05 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 201.399 uVs 16:49:06 Integrated Voltage Measurement... 16:49:14 Voltage Offset Correction: v_off_ini = 214.854 uV, v_off_fin = 200.780 uV 16:49:14 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -218.552 uVs 16:49:14 Integrated Voltage Measurement... 16:49:23 Voltage Offset Correction: v_off_ini = 209.225 uV, v_off_fin = 217.514 uV 16:49:23 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 211.093 uVs 16:49:28 I1Y Average and RMS of Measurements VT = -209.846 +- 3.654 uVs I1Y = -41.969 +- 0.731 uTm 16:49:30 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:49:33 Integrated Voltage Measurement... 16:49:42 Voltage Offset Correction: v_off_ini = 206.793 uV, v_off_fin = 193.624 uV 16:49:42 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -45.660 uVs 16:49:42 Integrated Voltage Measurement... 16:49:51 Voltage Offset Correction: v_off_ini = 209.706 uV, v_off_fin = 219.386 uV 16:49:51 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 56.649 uVs 16:49:51 Integrated Voltage Measurement... 16:50:00 Voltage Offset Correction: v_off_ini = 213.394 uV, v_off_fin = 200.334 uV 16:50:00 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -50.489 uVs 16:50:00 Integrated Voltage Measurement... 16:50:09 Voltage Offset Correction: v_off_ini = 209.395 uV, v_off_fin = 207.608 uV 16:50:09 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 54.943 uVs 16:50:09 Integrated Voltage Measurement... 16:50:18 Voltage Offset Correction: v_off_ini = 206.877 uV, v_off_fin = 195.154 uV 16:50:18 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -45.034 uVs 16:50:18 Integrated Voltage Measurement... 16:50:27 Voltage Offset Correction: v_off_ini = 218.612 uV, v_off_fin = 203.572 uV 16:50:27 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 39.548 uVs 16:50:29 I2X Average and RMS of Measurements VT = -48.720 +- 4.591 uVs I2X = 4.482 +- 0.422 uTm^2 16:50:32 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:50:34 Integrated Voltage Measurement... 16:50:43 Voltage Offset Correction: v_off_ini = 202.132 uV, v_off_fin = 207.989 uV 16:50:43 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -334.932 uVs 16:50:43 Integrated Voltage Measurement... 16:50:52 Voltage Offset Correction: v_off_ini = 212.296 uV, v_off_fin = 212.209 uV 16:50:52 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 317.592 uVs 16:50:52 Integrated Voltage Measurement... 16:51:01 Voltage Offset Correction: v_off_ini = 214.589 uV, v_off_fin = 212.054 uV 16:51:01 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -321.455 uVs 16:51:01 Integrated Voltage Measurement... 16:51:10 Voltage Offset Correction: v_off_ini = 227.441 uV, v_off_fin = 218.382 uV 16:51:10 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 310.368 uVs 16:51:10 Integrated Voltage Measurement... 16:51:19 Voltage Offset Correction: v_off_ini = 211.506 uV, v_off_fin = 217.316 uV 16:51:19 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = -332.241 uVs 16:51:19 Integrated Voltage Measurement... 16:51:28 Voltage Offset Correction: v_off_ini = 211.482 uV, v_off_fin = 224.742 uV 16:51:28 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 336.205 uVs 16:51:31 I2Y Average and RMS of Measurements VT = -325.465 +- 7.497 uVs I2Y = -29.943 +- 0.690 uTm^2 16:51:35 Wire Position: X0 = 0.0000 mm, Y0 = 0.5000 mm 16:51:39 Integrated Voltage Measurement... 16:51:48 Voltage Offset Correction: v_off_ini = 217.832 uV, v_off_fin = 205.454 uV 16:51:48 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -52.637 uVs 16:51:48 Integrated Voltage Measurement... 16:51:57 Voltage Offset Correction: v_off_ini = 220.124 uV, v_off_fin = 215.271 uV 16:51:57 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 43.439 uVs 16:51:57 Integrated Voltage Measurement... 16:52:06 Voltage Offset Correction: v_off_ini = 206.876 uV, v_off_fin = 217.564 uV 16:52:06 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -37.500 uVs 16:52:06 Integrated Voltage Measurement... 16:52:15 Voltage Offset Correction: v_off_ini = 228.052 uV, v_off_fin = 217.556 uV 16:52:15 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 42.045 uVs 16:52:15 Integrated Voltage Measurement... 16:52:24 Voltage Offset Correction: v_off_ini = 203.587 uV, v_off_fin = 202.174 uV 16:52:24 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -43.275 uVs 16:52:24 Integrated Voltage Measurement... 16:52:33 Voltage Offset Correction: v_off_ini = 210.446 uV, v_off_fin = 206.524 uV 16:52:33 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 51.594 uVs 16:52:37 I1X Average and RMS of Measurements VT = -45.082 +- 3.762 uVs I1X = 9.017 +- 0.753 uTm 16:52:39 Wire Position: X0 = 0.0000 mm, Y0 = 0.5000 mm 16:52:44 Integrated Voltage Measurement... 16:52:53 Voltage Offset Correction: v_off_ini = 208.150 uV, v_off_fin = 196.966 uV 16:52:53 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -213.632 uVs 16:52:53 Integrated Voltage Measurement... 16:53:02 Voltage Offset Correction: v_off_ini = 204.033 uV, v_off_fin = 205.906 uV 16:53:02 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 213.834 uVs 16:53:02 Integrated Voltage Measurement... 16:53:11 Voltage Offset Correction: v_off_ini = 205.948 uV, v_off_fin = 189.190 uV 16:53:11 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -198.118 uVs 16:53:11 Integrated Voltage Measurement... 16:53:20 Voltage Offset Correction: v_off_ini = 214.424 uV, v_off_fin = 224.535 uV 16:53:20 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 193.793 uVs 16:53:20 Integrated Voltage Measurement... 16:53:29 Voltage Offset Correction: v_off_ini = 208.005 uV, v_off_fin = 202.783 uV 16:53:29 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -199.364 uVs 16:53:29 Integrated Voltage Measurement... 16:53:38 Voltage Offset Correction: v_off_ini = 197.164 uV, v_off_fin = 214.144 uV 16:53:38 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 204.970 uVs 16:53:42 I1Y Average and RMS of Measurements VT = -203.952 +- 7.367 uVs I1Y = -40.790 +- 1.473 uTm 16:53:44 Wire Position: X0 = 0.0000 mm, Y0 = 0.5000 mm 16:53:47 Integrated Voltage Measurement... 16:53:56 Voltage Offset Correction: v_off_ini = 198.593 uV, v_off_fin = 196.833 uV 16:53:56 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -50.344 uVs 16:53:56 Integrated Voltage Measurement... 16:54:05 Voltage Offset Correction: v_off_ini = 216.804 uV, v_off_fin = 214.097 uV 16:54:05 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 49.668 uVs 16:54:05 Integrated Voltage Measurement... 16:54:14 Voltage Offset Correction: v_off_ini = 205.866 uV, v_off_fin = 203.437 uV 16:54:14 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -58.923 uVs 16:54:14 Integrated Voltage Measurement... 16:54:23 Voltage Offset Correction: v_off_ini = 219.267 uV, v_off_fin = 220.392 uV 16:54:23 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 71.149 uVs 16:54:23 Integrated Voltage Measurement... 16:54:32 Voltage Offset Correction: v_off_ini = 218.035 uV, v_off_fin = 213.091 uV 16:54:32 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -62.987 uVs 16:54:32 Integrated Voltage Measurement... 16:54:41 Voltage Offset Correction: v_off_ini = 212.083 uV, v_off_fin = 201.931 uV 16:54:41 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 62.631 uVs 16:54:43 I2X Average and RMS of Measurements VT = -59.284 +- 6.623 uVs I2X = 5.454 +- 0.610 uTm^2 16:54:46 Wire Position: X0 = 0.0000 mm, Y0 = 0.5000 mm 16:54:48 Integrated Voltage Measurement... 16:54:57 Voltage Offset Correction: v_off_ini = 209.238 uV, v_off_fin = 217.875 uV 16:54:57 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -332.428 uVs 16:54:57 Integrated Voltage Measurement... 16:55:06 Voltage Offset Correction: v_off_ini = 217.998 uV, v_off_fin = 219.439 uV 16:55:06 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 307.327 uVs 16:55:06 Integrated Voltage Measurement... 16:55:15 Voltage Offset Correction: v_off_ini = 201.643 uV, v_off_fin = 198.002 uV 16:55:15 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -320.702 uVs 16:55:15 Integrated Voltage Measurement... 16:55:24 Voltage Offset Correction: v_off_ini = 190.627 uV, v_off_fin = 201.403 uV 16:55:24 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 328.472 uVs 16:55:24 Integrated Voltage Measurement... 16:55:33 Voltage Offset Correction: v_off_ini = 203.810 uV, v_off_fin = 196.095 uV 16:55:33 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -336.300 uVs 16:55:33 Integrated Voltage Measurement... 16:55:42 Voltage Offset Correction: v_off_ini = 191.777 uV, v_off_fin = 197.404 uV 16:55:42 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 346.909 uVs 16:55:45 I2Y Average and RMS of Measurements VT = -328.690 +- 9.332 uVs I2Y = -30.239 +- 0.859 uTm^2 16:55:49 Wire Position: X0 = 0.0000 mm, Y0 = 1.0000 mm 16:55:53 Integrated Voltage Measurement... 16:56:02 Voltage Offset Correction: v_off_ini = 193.082 uV, v_off_fin = 182.779 uV 16:56:02 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -10.999 uVs 16:56:02 Integrated Voltage Measurement... 16:56:11 Voltage Offset Correction: v_off_ini = 196.981 uV, v_off_fin = 194.899 uV 16:56:11 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 56.304 uVs 16:56:11 Integrated Voltage Measurement... 16:56:20 Voltage Offset Correction: v_off_ini = 190.350 uV, v_off_fin = 189.043 uV 16:56:20 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -72.914 uVs 16:56:20 Integrated Voltage Measurement... 16:56:29 Voltage Offset Correction: v_off_ini = 183.640 uV, v_off_fin = 188.906 uV 16:56:29 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 65.953 uVs 16:56:29 Integrated Voltage Measurement... 16:56:38 Voltage Offset Correction: v_off_ini = 176.276 uV, v_off_fin = 175.242 uV 16:56:38 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -44.956 uVs 16:56:38 Integrated Voltage Measurement... 16:56:47 Voltage Offset Correction: v_off_ini = 165.703 uV, v_off_fin = 177.837 uV 16:56:47 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.4999 mm dXY = 0.4999 mm VT = 44.943 uVs 16:56:51 I1X Average and RMS of Measurements VT = -49.345 +- 14.935 uVs I1X = 9.869 +- 2.987 uTm 16:56:54 Wire Position: X0 = 0.0000 mm, Y0 = 1.0000 mm 16:56:58 Integrated Voltage Measurement... 16:57:07 Voltage Offset Correction: v_off_ini = 169.441 uV, v_off_fin = 162.925 uV 16:57:07 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -187.807 uVs 16:57:07 Integrated Voltage Measurement... 16:57:16 Voltage Offset Correction: v_off_ini = 183.848 uV, v_off_fin = 188.381 uV 16:57:16 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 198.318 uVs 16:57:16 Integrated Voltage Measurement... 16:57:25 Voltage Offset Correction: v_off_ini = 175.186 uV, v_off_fin = 165.798 uV 16:57:25 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -184.851 uVs 16:57:25 Integrated Voltage Measurement... 16:57:34 Voltage Offset Correction: v_off_ini = 170.832 uV, v_off_fin = 176.362 uV 16:57:34 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 185.717 uVs 16:57:34 Integrated Voltage Measurement... 16:57:43 Voltage Offset Correction: v_off_ini = 161.073 uV, v_off_fin = 155.545 uV 16:57:43 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -184.028 uVs 16:57:43 Integrated Voltage Measurement... 16:57:52 Voltage Offset Correction: v_off_ini = 164.994 uV, v_off_fin = 166.317 uV 16:57:52 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 187.319 uVs 16:57:56 I1Y Average and RMS of Measurements VT = -188.007 +- 3.579 uVs I1Y = -37.599 +- 0.715 uTm 16:57:58 Wire Position: X0 = 0.0000 mm, Y0 = 1.0000 mm 16:58:01 Integrated Voltage Measurement... 16:58:10 Voltage Offset Correction: v_off_ini = 175.273 uV, v_off_fin = 167.773 uV 16:58:10 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -70.783 uVs 16:58:10 Integrated Voltage Measurement... 16:58:19 Voltage Offset Correction: v_off_ini = 164.046 uV, v_off_fin = 161.582 uV 16:58:19 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 74.864 uVs 16:58:19 Integrated Voltage Measurement... 16:58:28 Voltage Offset Correction: v_off_ini = 163.026 uV, v_off_fin = 157.780 uV 16:58:28 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -69.290 uVs 16:58:28 Integrated Voltage Measurement... 16:58:37 Voltage Offset Correction: v_off_ini = 159.559 uV, v_off_fin = 186.435 uV 16:58:37 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 57.648 uVs 16:58:37 Integrated Voltage Measurement... 16:58:46 Voltage Offset Correction: v_off_ini = 178.585 uV, v_off_fin = 159.068 uV 16:58:46 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -50.087 uVs 16:58:46 Integrated Voltage Measurement... 16:58:55 Voltage Offset Correction: v_off_ini = 180.635 uV, v_off_fin = 181.592 uV 16:58:55 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 50.310 uVs 16:58:57 I2X Average and RMS of Measurements VT = -62.164 +- 9.283 uVs I2X = 5.719 +- 0.854 uTm^2 16:59:00 Wire Position: X0 = 0.0000 mm, Y0 = 1.0000 mm 16:59:03 Integrated Voltage Measurement... 16:59:11 Voltage Offset Correction: v_off_ini = 183.698 uV, v_off_fin = 181.703 uV 16:59:12 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -316.850 uVs 16:59:12 Integrated Voltage Measurement... 16:59:21 Voltage Offset Correction: v_off_ini = 173.895 uV, v_off_fin = 175.138 uV 16:59:21 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 321.828 uVs 16:59:21 Integrated Voltage Measurement... 16:59:30 Voltage Offset Correction: v_off_ini = 168.778 uV, v_off_fin = 183.061 uV 16:59:30 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -312.145 uVs 16:59:30 Integrated Voltage Measurement... 16:59:39 Voltage Offset Correction: v_off_ini = 179.821 uV, v_off_fin = 180.806 uV 16:59:39 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 305.106 uVs 16:59:39 Integrated Voltage Measurement... 16:59:48 Voltage Offset Correction: v_off_ini = 185.898 uV, v_off_fin = 183.826 uV 16:59:48 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -328.954 uVs 16:59:48 Integrated Voltage Measurement... 16:59:57 Voltage Offset Correction: v_off_ini = 172.888 uV, v_off_fin = 172.497 uV 16:59:57 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 333.482 uVs 16:59:59 I2Y Average and RMS of Measurements VT = -319.727 +- 9.227 uVs I2Y = -29.415 +- 0.849 uTm^2 17:00:15 Wire Position: X0 = -2.0000 mm, Y0 = 0.0000 mm 17:00:19 Integrated Voltage Measurement... 17:00:28 Voltage Offset Correction: v_off_ini = 180.096 uV, v_off_fin = 166.393 uV 17:00:28 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.4998 mm dXY = -0.4999 mm VT = -79.120 uVs 17:00:28 Integrated Voltage Measurement... 17:00:37 Voltage Offset Correction: v_off_ini = 172.314 uV, v_off_fin = 176.276 uV 17:00:37 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 79.820 uVs 17:00:37 Integrated Voltage Measurement... 17:00:46 Voltage Offset Correction: v_off_ini = 176.797 uV, v_off_fin = 166.969 uV 17:00:46 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -86.561 uVs 17:00:46 Integrated Voltage Measurement... 17:00:55 Voltage Offset Correction: v_off_ini = 163.296 uV, v_off_fin = 174.890 uV 17:00:55 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 80.742 uVs 17:00:55 Integrated Voltage Measurement... 17:01:04 Voltage Offset Correction: v_off_ini = 177.546 uV, v_off_fin = 179.763 uV 17:01:04 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -94.795 uVs 17:01:04 Integrated Voltage Measurement... 17:01:13 Voltage Offset Correction: v_off_ini = 179.522 uV, v_off_fin = 183.414 uV 17:01:13 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 87.143 uVs 17:01:17 I1X Average and RMS of Measurements VT = -84.697 +- 4.752 uVs I1X = 16.941 +- 0.950 uTm 17:01:20 Wire Position: X0 = -2.0000 mm, Y0 = 0.0000 mm 17:01:24 Integrated Voltage Measurement... 17:01:33 Voltage Offset Correction: v_off_ini = 183.592 uV, v_off_fin = 171.150 uV 17:01:33 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -201.538 uVs 17:01:33 Integrated Voltage Measurement... 17:01:42 Voltage Offset Correction: v_off_ini = 167.457 uV, v_off_fin = 188.712 uV 17:01:42 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 192.407 uVs 17:01:42 Integrated Voltage Measurement... 17:01:51 Voltage Offset Correction: v_off_ini = 183.609 uV, v_off_fin = 172.987 uV 17:01:51 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -192.172 uVs 17:01:51 Integrated Voltage Measurement... 17:02:00 Voltage Offset Correction: v_off_ini = 188.577 uV, v_off_fin = 195.231 uV 17:02:00 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 193.076 uVs 17:02:00 Integrated Voltage Measurement... 17:02:09 Voltage Offset Correction: v_off_ini = 190.034 uV, v_off_fin = 188.669 uV 17:02:09 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -205.069 uVs 17:02:09 Integrated Voltage Measurement... 17:02:18 Voltage Offset Correction: v_off_ini = 187.594 uV, v_off_fin = 197.197 uV 17:02:18 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 199.601 uVs 17:02:22 I1Y Average and RMS of Measurements VT = -197.311 +- 3.971 uVs I1Y = -39.462 +- 0.794 uTm 17:02:27 Wire Position: X0 = -2.0000 mm, Y0 = 0.0000 mm 17:02:29 Integrated Voltage Measurement... 17:02:38 Voltage Offset Correction: v_off_ini = 191.242 uV, v_off_fin = 185.810 uV 17:02:38 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -103.867 uVs 17:02:39 Integrated Voltage Measurement... 17:02:47 Voltage Offset Correction: v_off_ini = 208.765 uV, v_off_fin = 210.647 uV 17:02:47 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 96.078 uVs 17:02:48 Integrated Voltage Measurement... 17:02:56 Voltage Offset Correction: v_off_ini = 193.611 uV, v_off_fin = 185.259 uV 17:02:56 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -119.241 uVs 17:02:57 Integrated Voltage Measurement... 17:03:05 Voltage Offset Correction: v_off_ini = 183.570 uV, v_off_fin = 202.988 uV 17:03:06 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 117.995 uVs 17:03:06 Integrated Voltage Measurement... 17:03:14 Voltage Offset Correction: v_off_ini = 202.304 uV, v_off_fin = 181.723 uV 17:03:15 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -99.516 uVs 17:03:15 Integrated Voltage Measurement... 17:03:23 Voltage Offset Correction: v_off_ini = 181.236 uV, v_off_fin = 199.486 uV 17:03:24 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 98.113 uVs 17:03:26 I2X Average and RMS of Measurements VT = -105.802 +- 9.075 uVs I2X = 9.734 +- 0.835 uTm^2 17:03:31 Wire Position: X0 = -2.0000 mm, Y0 = 0.0000 mm 17:03:34 Integrated Voltage Measurement... 17:03:43 Voltage Offset Correction: v_off_ini = 211.889 uV, v_off_fin = 201.100 uV 17:03:43 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -323.060 uVs 17:03:43 Integrated Voltage Measurement... 17:03:52 Voltage Offset Correction: v_off_ini = 214.391 uV, v_off_fin = 188.683 uV 17:03:52 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 306.153 uVs 17:03:52 Integrated Voltage Measurement... 17:04:01 Voltage Offset Correction: v_off_ini = 184.389 uV, v_off_fin = 189.048 uV 17:04:01 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -302.131 uVs 17:04:01 Integrated Voltage Measurement... 17:04:10 Voltage Offset Correction: v_off_ini = 190.464 uV, v_off_fin = 195.602 uV 17:04:10 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 309.209 uVs 17:04:10 Integrated Voltage Measurement... 17:04:19 Voltage Offset Correction: v_off_ini = 196.250 uV, v_off_fin = 199.804 uV 17:04:19 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -308.009 uVs 17:04:19 Integrated Voltage Measurement... 17:04:28 Voltage Offset Correction: v_off_ini = 199.303 uV, v_off_fin = 199.060 uV 17:04:28 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 310.886 uVs 17:04:30 I2Y Average and RMS of Measurements VT = -309.908 +- 3.663 uVs I2Y = -28.512 +- 0.337 uTm^2 17:04:35 Wire Position: X0 = -1.0000 mm, Y0 = 0.0000 mm 17:04:39 Integrated Voltage Measurement... 17:04:48 Voltage Offset Correction: v_off_ini = 193.358 uV, v_off_fin = 187.996 uV 17:04:48 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -76.718 uVs 17:04:48 Integrated Voltage Measurement... 17:04:57 Voltage Offset Correction: v_off_ini = 196.137 uV, v_off_fin = 189.986 uV 17:04:57 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 71.080 uVs 17:04:57 Integrated Voltage Measurement... 17:05:06 Voltage Offset Correction: v_off_ini = 172.448 uV, v_off_fin = 171.084 uV 17:05:06 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -73.914 uVs 17:05:06 Integrated Voltage Measurement... 17:05:15 Voltage Offset Correction: v_off_ini = 188.817 uV, v_off_fin = 198.832 uV 17:05:15 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 61.220 uVs 17:05:15 Integrated Voltage Measurement... 17:05:24 Voltage Offset Correction: v_off_ini = 191.960 uV, v_off_fin = 187.316 uV 17:05:24 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -94.037 uVs 17:05:24 Integrated Voltage Measurement... 17:05:33 Voltage Offset Correction: v_off_ini = 185.250 uV, v_off_fin = 181.862 uV 17:05:33 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 76.924 uVs 17:05:37 I1X Average and RMS of Measurements VT = -75.649 +- 7.417 uVs I1X = 15.130 +- 1.483 uTm 17:05:39 Wire Position: X0 = -1.0000 mm, Y0 = 0.0000 mm 17:05:44 Integrated Voltage Measurement... 17:05:53 Voltage Offset Correction: v_off_ini = 185.501 uV, v_off_fin = 177.535 uV 17:05:53 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -212.838 uVs 17:05:53 Integrated Voltage Measurement... 17:06:02 Voltage Offset Correction: v_off_ini = 189.063 uV, v_off_fin = 205.769 uV 17:06:02 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 195.230 uVs 17:06:02 Integrated Voltage Measurement... 17:06:11 Voltage Offset Correction: v_off_ini = 187.617 uV, v_off_fin = 173.034 uV 17:06:11 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -210.268 uVs 17:06:11 Integrated Voltage Measurement... 17:06:20 Voltage Offset Correction: v_off_ini = 193.843 uV, v_off_fin = 188.528 uV 17:06:20 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 195.347 uVs 17:06:20 Integrated Voltage Measurement... 17:06:29 Voltage Offset Correction: v_off_ini = 186.638 uV, v_off_fin = 191.513 uV 17:06:29 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -206.208 uVs 17:06:29 Integrated Voltage Measurement... 17:06:37 Voltage Offset Correction: v_off_ini = 192.414 uV, v_off_fin = 207.276 uV 17:06:38 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 196.524 uVs 17:06:42 I1Y Average and RMS of Measurements VT = -202.736 +- 1.090 uVs I1Y = -40.547 +- 0.218 uTm 17:06:47 Wire Position: X0 = -1.0000 mm, Y0 = 0.0000 mm 17:06:49 Integrated Voltage Measurement... 17:06:58 Voltage Offset Correction: v_off_ini = 192.963 uV, v_off_fin = 181.347 uV 17:06:58 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -83.500 uVs 17:06:58 Integrated Voltage Measurement... 17:07:07 Voltage Offset Correction: v_off_ini = 206.043 uV, v_off_fin = 205.248 uV 17:07:07 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 83.602 uVs 17:07:07 Integrated Voltage Measurement... 17:07:16 Voltage Offset Correction: v_off_ini = 214.107 uV, v_off_fin = 209.359 uV 17:07:16 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -75.900 uVs 17:07:16 Integrated Voltage Measurement... 17:07:25 Voltage Offset Correction: v_off_ini = 226.270 uV, v_off_fin = 231.400 uV 17:07:25 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 80.813 uVs 17:07:25 Integrated Voltage Measurement... 17:07:34 Voltage Offset Correction: v_off_ini = 229.490 uV, v_off_fin = 228.619 uV 17:07:34 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -67.710 uVs 17:07:34 Integrated Voltage Measurement... 17:07:43 Voltage Offset Correction: v_off_ini = 235.670 uV, v_off_fin = 240.802 uV 17:07:43 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 97.670 uVs 17:07:46 I2X Average and RMS of Measurements VT = -81.533 +- 2.273 uVs I2X = 7.501 +- 0.209 uTm^2 17:07:51 Wire Position: X0 = -1.0000 mm, Y0 = 0.0000 mm 17:07:53 Integrated Voltage Measurement... 17:08:02 Voltage Offset Correction: v_off_ini = 220.171 uV, v_off_fin = 221.822 uV 17:08:02 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -312.858 uVs 17:08:02 Integrated Voltage Measurement... 17:08:11 Voltage Offset Correction: v_off_ini = 225.135 uV, v_off_fin = 231.154 uV 17:08:11 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 316.212 uVs 17:08:11 Integrated Voltage Measurement... 17:08:20 Voltage Offset Correction: v_off_ini = 213.618 uV, v_off_fin = 213.203 uV 17:08:20 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -315.559 uVs 17:08:20 Integrated Voltage Measurement... 17:08:29 Voltage Offset Correction: v_off_ini = 209.726 uV, v_off_fin = 217.276 uV 17:08:29 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 326.499 uVs 17:08:30 Integrated Voltage Measurement... 17:08:38 Voltage Offset Correction: v_off_ini = 215.551 uV, v_off_fin = 218.274 uV 17:08:38 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -313.736 uVs 17:08:39 Integrated Voltage Measurement... 17:08:47 Voltage Offset Correction: v_off_ini = 219.742 uV, v_off_fin = 228.095 uV 17:08:48 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 312.302 uVs 17:08:50 I2Y Average and RMS of Measurements VT = -316.194 +- 3.474 uVs I2Y = -29.090 +- 0.320 uTm^2 17:08:54 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 17:08:59 Integrated Voltage Measurement... 17:09:08 Voltage Offset Correction: v_off_ini = 224.048 uV, v_off_fin = 211.348 uV 17:09:08 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -68.869 uVs 17:09:08 Integrated Voltage Measurement... 17:09:17 Voltage Offset Correction: v_off_ini = 223.265 uV, v_off_fin = 215.496 uV 17:09:17 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 65.146 uVs 17:09:17 Integrated Voltage Measurement... 17:09:26 Voltage Offset Correction: v_off_ini = 213.985 uV, v_off_fin = 204.371 uV 17:09:26 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -51.977 uVs 17:09:26 Integrated Voltage Measurement... 17:09:35 Voltage Offset Correction: v_off_ini = 218.531 uV, v_off_fin = 235.712 uV 17:09:35 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 69.205 uVs 17:09:35 Integrated Voltage Measurement... 17:09:44 Voltage Offset Correction: v_off_ini = 230.042 uV, v_off_fin = 236.503 uV 17:09:44 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -73.742 uVs 17:09:44 Integrated Voltage Measurement... 17:09:53 Voltage Offset Correction: v_off_ini = 235.455 uV, v_off_fin = 234.456 uV 17:09:53 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 71.230 uVs 17:09:57 I1X Average and RMS of Measurements VT = -66.695 +- 4.861 uVs I1X = 13.339 +- 0.972 uTm 17:09:59 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 17:10:04 Integrated Voltage Measurement... 17:10:13 Voltage Offset Correction: v_off_ini = 241.621 uV, v_off_fin = 216.982 uV 17:10:13 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -206.396 uVs 17:10:13 Integrated Voltage Measurement... 17:10:21 Voltage Offset Correction: v_off_ini = 227.986 uV, v_off_fin = 233.249 uV 17:10:22 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 207.861 uVs 17:10:22 Integrated Voltage Measurement... 17:10:30 Voltage Offset Correction: v_off_ini = 221.370 uV, v_off_fin = 204.271 uV 17:10:31 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -205.256 uVs 17:10:31 Integrated Voltage Measurement... 17:10:39 Voltage Offset Correction: v_off_ini = 217.820 uV, v_off_fin = 233.170 uV 17:10:39 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 184.808 uVs 17:10:39 Integrated Voltage Measurement... 17:10:48 Voltage Offset Correction: v_off_ini = 212.143 uV, v_off_fin = 210.233 uV 17:10:48 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -190.766 uVs 17:10:48 Integrated Voltage Measurement... 17:10:57 Voltage Offset Correction: v_off_ini = 214.973 uV, v_off_fin = 230.443 uV 17:10:57 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 197.511 uVs 17:11:02 I1Y Average and RMS of Measurements VT = -198.766 +- 5.924 uVs I1Y = -39.755 +- 1.185 uTm 17:11:07 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 17:11:09 Integrated Voltage Measurement... 17:11:18 Voltage Offset Correction: v_off_ini = 210.969 uV, v_off_fin = 218.212 uV 17:11:18 Integrated Voltage Measurement dXYus = -1.0001 mm, dXYds = 0.0000 mm dXY = -1.0001 mm VT = -75.269 uVs 17:11:18 Integrated Voltage Measurement... 17:11:27 Voltage Offset Correction: v_off_ini = 232.188 uV, v_off_fin = 227.557 uV 17:11:27 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 70.739 uVs 17:11:27 Integrated Voltage Measurement... 17:11:36 Voltage Offset Correction: v_off_ini = 236.154 uV, v_off_fin = 223.724 uV 17:11:36 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -90.213 uVs 17:11:36 Integrated Voltage Measurement... 17:11:45 Voltage Offset Correction: v_off_ini = 227.128 uV, v_off_fin = 226.001 uV 17:11:45 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 84.298 uVs 17:11:45 Integrated Voltage Measurement... 17:11:54 Voltage Offset Correction: v_off_ini = 216.665 uV, v_off_fin = 203.834 uV 17:11:54 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -70.986 uVs 17:11:54 Integrated Voltage Measurement... 17:12:03 Voltage Offset Correction: v_off_ini = 222.354 uV, v_off_fin = 237.679 uV 17:12:03 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 83.203 uVs 17:12:06 I2X Average and RMS of Measurements VT = -79.118 +- 5.992 uVs I2X = 7.279 +- 0.551 uTm^2 17:12:11 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 17:12:13 Integrated Voltage Measurement... 17:12:22 Voltage Offset Correction: v_off_ini = 230.505 uV, v_off_fin = 229.987 uV 17:12:22 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -325.425 uVs 17:12:22 Integrated Voltage Measurement... 17:12:31 Voltage Offset Correction: v_off_ini = 228.772 uV, v_off_fin = 226.159 uV 17:12:31 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 320.253 uVs 17:12:31 Integrated Voltage Measurement... 17:12:40 Voltage Offset Correction: v_off_ini = 222.863 uV, v_off_fin = 221.407 uV 17:12:40 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -307.019 uVs 17:12:40 Integrated Voltage Measurement... 17:12:49 Voltage Offset Correction: v_off_ini = 229.493 uV, v_off_fin = 228.878 uV 17:12:49 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 314.416 uVs 17:12:49 Integrated Voltage Measurement... 17:12:58 Voltage Offset Correction: v_off_ini = 222.719 uV, v_off_fin = 221.793 uV 17:12:58 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -314.048 uVs 17:12:58 Integrated Voltage Measurement... 17:13:07 Voltage Offset Correction: v_off_ini = 232.389 uV, v_off_fin = 227.753 uV 17:13:07 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 321.546 uVs 17:13:10 I2Y Average and RMS of Measurements VT = -317.118 +- 4.972 uVs I2Y = -29.175 +- 0.457 uTm^2 17:13:14 Wire Position: X0 = 1.0000 mm, Y0 = 0.0000 mm 17:13:19 Integrated Voltage Measurement... 17:13:28 Voltage Offset Correction: v_off_ini = 238.078 uV, v_off_fin = 232.443 uV 17:13:28 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4998 mm dXY = -0.4999 mm VT = -70.241 uVs 17:13:28 Integrated Voltage Measurement... 17:13:37 Voltage Offset Correction: v_off_ini = 225.741 uV, v_off_fin = 233.792 uV 17:13:37 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 40.400 uVs 17:13:37 Integrated Voltage Measurement... 17:13:46 Voltage Offset Correction: v_off_ini = 231.354 uV, v_off_fin = 225.283 uV 17:13:46 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -59.629 uVs 17:13:46 Integrated Voltage Measurement... 17:13:55 Voltage Offset Correction: v_off_ini = 227.623 uV, v_off_fin = 241.458 uV 17:13:55 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 57.851 uVs 17:13:55 Integrated Voltage Measurement... 17:14:04 Voltage Offset Correction: v_off_ini = 232.710 uV, v_off_fin = 238.131 uV 17:14:04 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -68.106 uVs 17:14:04 Integrated Voltage Measurement... 17:14:13 Voltage Offset Correction: v_off_ini = 246.318 uV, v_off_fin = 250.429 uV 17:14:13 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 58.600 uVs 17:14:17 I1X Average and RMS of Measurements VT = -59.138 +- 3.291 uVs I1X = 11.828 +- 0.658 uTm 17:14:19 Wire Position: X0 = 1.0000 mm, Y0 = 0.0000 mm 17:14:23 Integrated Voltage Measurement... 17:14:32 Voltage Offset Correction: v_off_ini = 242.789 uV, v_off_fin = 237.252 uV 17:14:32 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -209.879 uVs 17:14:32 Integrated Voltage Measurement... 17:14:41 Voltage Offset Correction: v_off_ini = 245.397 uV, v_off_fin = 244.213 uV 17:14:41 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 220.954 uVs 17:14:41 Integrated Voltage Measurement... 17:14:50 Voltage Offset Correction: v_off_ini = 239.184 uV, v_off_fin = 230.710 uV 17:14:50 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -190.715 uVs 17:14:50 Integrated Voltage Measurement... 17:14:59 Voltage Offset Correction: v_off_ini = 255.517 uV, v_off_fin = 258.277 uV 17:14:59 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 214.231 uVs 17:14:59 Integrated Voltage Measurement... 17:15:08 Voltage Offset Correction: v_off_ini = 245.373 uV, v_off_fin = 245.344 uV 17:15:08 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -194.178 uVs 17:15:08 Integrated Voltage Measurement... 17:15:17 Voltage Offset Correction: v_off_ini = 252.050 uV, v_off_fin = 267.034 uV 17:15:17 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 196.527 uVs 17:15:22 I1Y Average and RMS of Measurements VT = -204.414 +- 8.305 uVs I1Y = -40.883 +- 1.661 uTm 17:15:27 Wire Position: X0 = 1.0000 mm, Y0 = 0.0000 mm 17:15:29 Integrated Voltage Measurement... 17:15:38 Voltage Offset Correction: v_off_ini = 246.913 uV, v_off_fin = 249.126 uV 17:15:38 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -42.503 uVs 17:15:38 Integrated Voltage Measurement... 17:15:47 Voltage Offset Correction: v_off_ini = 248.709 uV, v_off_fin = 257.879 uV 17:15:47 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 80.130 uVs 17:15:47 Integrated Voltage Measurement... 17:15:56 Voltage Offset Correction: v_off_ini = 250.736 uV, v_off_fin = 239.431 uV 17:15:56 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -59.221 uVs 17:15:56 Integrated Voltage Measurement... 17:16:05 Voltage Offset Correction: v_off_ini = 240.655 uV, v_off_fin = 257.951 uV 17:16:05 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 58.930 uVs 17:16:05 Integrated Voltage Measurement... 17:16:14 Voltage Offset Correction: v_off_ini = 243.401 uV, v_off_fin = 233.605 uV 17:16:14 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -82.864 uVs 17:16:14 Integrated Voltage Measurement... 17:16:23 Voltage Offset Correction: v_off_ini = 245.498 uV, v_off_fin = 248.119 uV 17:16:23 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 63.407 uVs 17:16:26 I2X Average and RMS of Measurements VT = -64.509 +- 6.168 uVs I2X = 5.935 +- 0.567 uTm^2 17:16:31 Wire Position: X0 = 1.0000 mm, Y0 = 0.0000 mm 17:16:33 Integrated Voltage Measurement... 17:16:42 Voltage Offset Correction: v_off_ini = 231.403 uV, v_off_fin = 240.160 uV 17:16:42 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -304.525 uVs 17:16:42 Integrated Voltage Measurement... 17:16:51 Voltage Offset Correction: v_off_ini = 235.523 uV, v_off_fin = 230.328 uV 17:16:51 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 305.739 uVs 17:16:51 Integrated Voltage Measurement... 17:17:00 Voltage Offset Correction: v_off_ini = 231.725 uV, v_off_fin = 227.393 uV 17:17:00 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -302.524 uVs 17:17:00 Integrated Voltage Measurement... 17:17:09 Voltage Offset Correction: v_off_ini = 224.188 uV, v_off_fin = 232.349 uV 17:17:09 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 305.833 uVs 17:17:09 Integrated Voltage Measurement... 17:17:18 Voltage Offset Correction: v_off_ini = 218.415 uV, v_off_fin = 230.190 uV 17:17:18 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -319.092 uVs 17:17:18 Integrated Voltage Measurement... 17:17:27 Voltage Offset Correction: v_off_ini = 233.284 uV, v_off_fin = 232.077 uV 17:17:27 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 320.714 uVs 17:17:30 I2Y Average and RMS of Measurements VT = -309.738 +- 7.198 uVs I2Y = -28.496 +- 0.662 uTm^2 17:17:34 Wire Position: X0 = 2.0000 mm, Y0 = 0.0000 mm 17:17:39 Integrated Voltage Measurement... 17:17:48 Voltage Offset Correction: v_off_ini = 218.008 uV, v_off_fin = 214.158 uV 17:17:48 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -51.563 uVs 17:17:48 Integrated Voltage Measurement... 17:17:57 Voltage Offset Correction: v_off_ini = 213.845 uV, v_off_fin = 209.663 uV 17:17:57 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 52.931 uVs 17:17:57 Integrated Voltage Measurement... 17:18:06 Voltage Offset Correction: v_off_ini = 204.309 uV, v_off_fin = 199.399 uV 17:18:06 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -44.323 uVs 17:18:06 Integrated Voltage Measurement... 17:18:15 Voltage Offset Correction: v_off_ini = 221.166 uV, v_off_fin = 216.447 uV 17:18:15 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = 41.800 uVs 17:18:15 Integrated Voltage Measurement... 17:18:23 Voltage Offset Correction: v_off_ini = 215.833 uV, v_off_fin = 207.886 uV 17:18:24 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -75.048 uVs 17:18:24 Integrated Voltage Measurement... 17:18:32 Voltage Offset Correction: v_off_ini = 204.020 uV, v_off_fin = 219.359 uV 17:18:33 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 51.816 uVs 17:18:37 I1X Average and RMS of Measurements VT = -52.913 +- 8.330 uVs I1X = 10.583 +- 1.666 uTm 17:18:39 Wire Position: X0 = 2.0000 mm, Y0 = 0.0000 mm 17:18:43 Integrated Voltage Measurement... 17:18:52 Voltage Offset Correction: v_off_ini = 219.570 uV, v_off_fin = 217.306 uV 17:18:52 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -190.843 uVs 17:18:52 Integrated Voltage Measurement... 17:19:01 Voltage Offset Correction: v_off_ini = 205.858 uV, v_off_fin = 210.683 uV 17:19:01 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 200.640 uVs 17:19:01 Integrated Voltage Measurement... 17:19:10 Voltage Offset Correction: v_off_ini = 203.197 uV, v_off_fin = 189.544 uV 17:19:10 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -206.647 uVs 17:19:10 Integrated Voltage Measurement... 17:19:19 Voltage Offset Correction: v_off_ini = 190.850 uV, v_off_fin = 219.120 uV 17:19:19 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 203.093 uVs 17:19:19 Integrated Voltage Measurement... 17:19:28 Voltage Offset Correction: v_off_ini = 208.836 uV, v_off_fin = 204.791 uV 17:19:28 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.5000 mm dXY = 0.4999 mm VT = -207.619 uVs 17:19:28 Integrated Voltage Measurement... 17:19:37 Voltage Offset Correction: v_off_ini = 213.803 uV, v_off_fin = 218.096 uV 17:19:37 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 188.103 uVs 17:19:41 I1Y Average and RMS of Measurements VT = -199.491 +- 3.901 uVs I1Y = -39.899 +- 0.780 uTm 17:19:47 Wire Position: X0 = 2.0000 mm, Y0 = 0.0000 mm 17:19:49 Integrated Voltage Measurement... 17:19:58 Voltage Offset Correction: v_off_ini = 199.649 uV, v_off_fin = 206.287 uV 17:19:58 Integrated Voltage Measurement dXYus = -1.0001 mm, dXYds = 0.0000 mm dXY = -1.0001 mm VT = -59.389 uVs 17:19:58 Integrated Voltage Measurement... 17:20:07 Voltage Offset Correction: v_off_ini = 207.113 uV, v_off_fin = 209.939 uV 17:20:07 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 48.644 uVs 17:20:07 Integrated Voltage Measurement... 17:20:16 Voltage Offset Correction: v_off_ini = 213.840 uV, v_off_fin = 199.906 uV 17:20:16 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -54.220 uVs 17:20:16 Integrated Voltage Measurement... 17:20:25 Voltage Offset Correction: v_off_ini = 216.318 uV, v_off_fin = 216.004 uV 17:20:25 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 73.533 uVs 17:20:25 Integrated Voltage Measurement... 17:20:34 Voltage Offset Correction: v_off_ini = 207.251 uV, v_off_fin = 191.597 uV 17:20:34 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -61.173 uVs 17:20:34 Integrated Voltage Measurement... 17:20:43 Voltage Offset Correction: v_off_ini = 196.795 uV, v_off_fin = 216.576 uV 17:20:43 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 73.063 uVs 17:20:46 I2X Average and RMS of Measurements VT = -61.670 +- 5.571 uVs I2X = 5.674 +- 0.513 uTm^2 17:20:51 Wire Position: X0 = 2.0000 mm, Y0 = 0.0000 mm 17:20:53 Integrated Voltage Measurement... 17:21:02 Voltage Offset Correction: v_off_ini = 212.192 uV, v_off_fin = 211.415 uV 17:21:02 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -313.802 uVs 17:21:02 Integrated Voltage Measurement... 17:21:11 Voltage Offset Correction: v_off_ini = 205.103 uV, v_off_fin = 205.930 uV 17:21:11 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 308.477 uVs 17:21:11 Integrated Voltage Measurement... 17:21:20 Voltage Offset Correction: v_off_ini = 205.584 uV, v_off_fin = 212.900 uV 17:21:20 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -320.332 uVs 17:21:20 Integrated Voltage Measurement... 17:21:29 Voltage Offset Correction: v_off_ini = 216.981 uV, v_off_fin = 213.780 uV 17:21:29 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 305.061 uVs 17:21:29 Integrated Voltage Measurement... 17:21:38 Voltage Offset Correction: v_off_ini = 223.769 uV, v_off_fin = 209.900 uV 17:21:38 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -324.947 uVs 17:21:38 Integrated Voltage Measurement... 17:21:47 Voltage Offset Correction: v_off_ini = 189.810 uV, v_off_fin = 199.982 uV 17:21:47 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 307.634 uVs 17:21:50 I2Y Average and RMS of Measurements VT = -313.375 +- 2.157 uVs I2Y = -28.831 +- 0.198 uTm^2 17:21:57 Wire Position: X0 = 0.0000 mm, Y0 = -1.0000 mm 17:22:02 Integrated Voltage Measurement... 17:22:11 Voltage Offset Correction: v_off_ini = 215.048 uV, v_off_fin = 210.307 uV 17:22:11 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -58.381 uVs 17:22:11 Integrated Voltage Measurement... 17:22:19 Voltage Offset Correction: v_off_ini = 208.237 uV, v_off_fin = 216.185 uV 17:22:20 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 76.128 uVs 17:22:20 Integrated Voltage Measurement... 17:22:28 Voltage Offset Correction: v_off_ini = 217.329 uV, v_off_fin = 218.893 uV 17:22:29 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -49.565 uVs 17:22:29 Integrated Voltage Measurement... 17:22:37 Voltage Offset Correction: v_off_ini = 198.602 uV, v_off_fin = 217.640 uV 17:22:38 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 61.986 uVs 17:22:38 Integrated Voltage Measurement... 17:22:46 Voltage Offset Correction: v_off_ini = 221.875 uV, v_off_fin = 206.963 uV 17:22:46 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.4999 mm dXY = -0.4999 mm VT = -49.366 uVs 17:22:47 Integrated Voltage Measurement... 17:22:55 Voltage Offset Correction: v_off_ini = 212.436 uV, v_off_fin = 221.073 uV 17:22:55 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = 46.349 uVs 17:23:00 I1X Average and RMS of Measurements VT = -56.962 +- 7.963 uVs I1X = 11.393 +- 1.592 uTm 17:23:02 Wire Position: X0 = 0.0000 mm, Y0 = -1.0000 mm 17:23:06 Integrated Voltage Measurement... 17:23:15 Voltage Offset Correction: v_off_ini = 213.345 uV, v_off_fin = 199.285 uV 17:23:15 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -209.956 uVs 17:23:15 Integrated Voltage Measurement... 17:23:24 Voltage Offset Correction: v_off_ini = 203.965 uV, v_off_fin = 223.189 uV 17:23:24 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 208.564 uVs 17:23:24 Integrated Voltage Measurement... 17:23:33 Voltage Offset Correction: v_off_ini = 203.387 uV, v_off_fin = 201.083 uV 17:23:33 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -194.416 uVs 17:23:33 Integrated Voltage Measurement... 17:23:42 Voltage Offset Correction: v_off_ini = 213.393 uV, v_off_fin = 227.280 uV 17:23:42 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 209.043 uVs 17:23:42 Integrated Voltage Measurement... 17:23:51 Voltage Offset Correction: v_off_ini = 213.384 uV, v_off_fin = 202.074 uV 17:23:51 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -203.239 uVs 17:23:51 Integrated Voltage Measurement... 17:24:00 Voltage Offset Correction: v_off_ini = 212.111 uV, v_off_fin = 221.182 uV 17:24:00 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 196.939 uVs 17:24:04 I1Y Average and RMS of Measurements VT = -203.693 +- 3.993 uVs I1Y = -40.739 +- 0.799 uTm 17:24:07 Wire Position: X0 = 0.0000 mm, Y0 = -1.0000 mm 17:24:09 Integrated Voltage Measurement... 17:24:18 Voltage Offset Correction: v_off_ini = 203.627 uV, v_off_fin = 198.787 uV 17:24:18 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -34.778 uVs 17:24:19 Integrated Voltage Measurement... 17:24:27 Voltage Offset Correction: v_off_ini = 217.285 uV, v_off_fin = 219.171 uV 17:24:27 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 55.775 uVs 17:24:28 Integrated Voltage Measurement... 17:24:36 Voltage Offset Correction: v_off_ini = 218.523 uV, v_off_fin = 213.714 uV 17:24:36 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -54.337 uVs 17:24:37 Integrated Voltage Measurement... 17:24:45 Voltage Offset Correction: v_off_ini = 217.573 uV, v_off_fin = 224.788 uV 17:24:45 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 67.685 uVs 17:24:46 Integrated Voltage Measurement... 17:24:54 Voltage Offset Correction: v_off_ini = 205.968 uV, v_off_fin = 217.135 uV 17:24:55 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -73.189 uVs 17:24:55 Integrated Voltage Measurement... 17:25:03 Voltage Offset Correction: v_off_ini = 217.406 uV, v_off_fin = 229.771 uV 17:25:04 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 50.699 uVs 17:25:06 I2X Average and RMS of Measurements VT = -56.077 +- 7.647 uVs I2X = 5.159 +- 0.704 uTm^2 17:25:08 Wire Position: X0 = 0.0000 mm, Y0 = -1.0000 mm 17:25:11 Integrated Voltage Measurement... 17:25:20 Voltage Offset Correction: v_off_ini = 212.162 uV, v_off_fin = 211.263 uV 17:25:20 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -316.649 uVs 17:25:20 Integrated Voltage Measurement... 17:25:29 Voltage Offset Correction: v_off_ini = 215.722 uV, v_off_fin = 220.361 uV 17:25:29 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 331.140 uVs 17:25:29 Integrated Voltage Measurement... 17:25:38 Voltage Offset Correction: v_off_ini = 218.150 uV, v_off_fin = 210.694 uV 17:25:38 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -340.023 uVs 17:25:38 Integrated Voltage Measurement... 17:25:47 Voltage Offset Correction: v_off_ini = 218.574 uV, v_off_fin = 218.734 uV 17:25:47 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 329.281 uVs 17:25:47 Integrated Voltage Measurement... 17:25:56 Voltage Offset Correction: v_off_ini = 207.869 uV, v_off_fin = 215.655 uV 17:25:56 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -320.846 uVs 17:25:56 Integrated Voltage Measurement... 17:26:05 Voltage Offset Correction: v_off_ini = 225.537 uV, v_off_fin = 211.827 uV 17:26:05 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 316.832 uVs 17:26:07 I2Y Average and RMS of Measurements VT = -325.795 +- 6.594 uVs I2Y = -29.973 +- 0.607 uTm^2 17:26:11 Wire Position: X0 = 0.0000 mm, Y0 = -0.5000 mm 17:26:16 Integrated Voltage Measurement... 17:26:25 Voltage Offset Correction: v_off_ini = 209.846 uV, v_off_fin = 226.204 uV 17:26:25 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.4998 mm dXY = -0.4999 mm VT = -74.397 uVs 17:26:25 Integrated Voltage Measurement... 17:26:34 Voltage Offset Correction: v_off_ini = 215.844 uV, v_off_fin = 220.684 uV 17:26:34 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 41.880 uVs 17:26:34 Integrated Voltage Measurement... 17:26:43 Voltage Offset Correction: v_off_ini = 219.128 uV, v_off_fin = 202.993 uV 17:26:43 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -67.273 uVs 17:26:43 Integrated Voltage Measurement... 17:26:52 Voltage Offset Correction: v_off_ini = 211.652 uV, v_off_fin = 216.018 uV 17:26:52 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.4999 mm dXY = 0.4999 mm VT = 68.685 uVs 17:26:52 Integrated Voltage Measurement... 17:27:01 Voltage Offset Correction: v_off_ini = 206.380 uV, v_off_fin = 200.747 uV 17:27:01 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -57.352 uVs 17:27:01 Integrated Voltage Measurement... 17:27:10 Voltage Offset Correction: v_off_ini = 208.385 uV, v_off_fin = 217.438 uV 17:27:10 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 66.910 uVs 17:27:14 I1X Average and RMS of Measurements VT = -62.750 +- 4.041 uVs I1X = 12.551 +- 0.808 uTm 17:27:16 Wire Position: X0 = 0.0000 mm, Y0 = -0.5000 mm 17:27:21 Integrated Voltage Measurement... 17:27:29 Voltage Offset Correction: v_off_ini = 206.033 uV, v_off_fin = 194.199 uV 17:27:30 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -204.811 uVs 17:27:30 Integrated Voltage Measurement... 17:27:38 Voltage Offset Correction: v_off_ini = 199.762 uV, v_off_fin = 212.795 uV 17:27:38 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 193.491 uVs 17:27:39 Integrated Voltage Measurement... 17:27:47 Voltage Offset Correction: v_off_ini = 201.292 uV, v_off_fin = 199.380 uV 17:27:47 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -201.237 uVs 17:27:47 Integrated Voltage Measurement... 17:27:56 Voltage Offset Correction: v_off_ini = 201.329 uV, v_off_fin = 202.512 uV 17:27:56 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 214.072 uVs 17:27:56 Integrated Voltage Measurement... 17:28:05 Voltage Offset Correction: v_off_ini = 195.591 uV, v_off_fin = 191.020 uV 17:28:05 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -204.298 uVs 17:28:05 Integrated Voltage Measurement... 17:28:14 Voltage Offset Correction: v_off_ini = 202.339 uV, v_off_fin = 221.077 uV 17:28:14 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 191.693 uVs 17:28:19 I1Y Average and RMS of Measurements VT = -201.600 +- 4.307 uVs I1Y = -40.320 +- 0.861 uTm 17:28:21 Wire Position: X0 = 0.0000 mm, Y0 = -0.5000 mm 17:28:24 Integrated Voltage Measurement... 17:28:33 Voltage Offset Correction: v_off_ini = 208.626 uV, v_off_fin = 199.828 uV 17:28:33 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -70.753 uVs 17:28:33 Integrated Voltage Measurement... 17:28:42 Voltage Offset Correction: v_off_ini = 183.725 uV, v_off_fin = 198.534 uV 17:28:42 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 74.362 uVs 17:28:42 Integrated Voltage Measurement... 17:28:51 Voltage Offset Correction: v_off_ini = 193.619 uV, v_off_fin = 184.412 uV 17:28:51 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -70.857 uVs 17:28:51 Integrated Voltage Measurement... 17:29:00 Voltage Offset Correction: v_off_ini = 193.858 uV, v_off_fin = 192.801 uV 17:29:00 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 64.661 uVs 17:29:00 Integrated Voltage Measurement... 17:29:09 Voltage Offset Correction: v_off_ini = 202.251 uV, v_off_fin = 181.494 uV 17:29:09 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -96.735 uVs 17:29:09 Integrated Voltage Measurement... 17:29:18 Voltage Offset Correction: v_off_ini = 193.524 uV, v_off_fin = 192.525 uV 17:29:18 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 61.229 uVs 17:29:20 I2X Average and RMS of Measurements VT = -73.099 +- 4.598 uVs I2X = 6.725 +- 0.423 uTm^2 17:29:23 Wire Position: X0 = 0.0000 mm, Y0 = -0.5000 mm 17:29:25 Integrated Voltage Measurement... 17:29:34 Voltage Offset Correction: v_off_ini = 195.262 uV, v_off_fin = 204.022 uV 17:29:34 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -300.928 uVs 17:29:34 Integrated Voltage Measurement... 17:29:43 Voltage Offset Correction: v_off_ini = 199.454 uV, v_off_fin = 204.271 uV 17:29:43 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = 317.551 uVs 17:29:43 Integrated Voltage Measurement... 17:29:52 Voltage Offset Correction: v_off_ini = 215.261 uV, v_off_fin = 214.344 uV 17:29:52 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -327.410 uVs 17:29:52 Integrated Voltage Measurement... 17:30:01 Voltage Offset Correction: v_off_ini = 205.419 uV, v_off_fin = 194.292 uV 17:30:01 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 324.869 uVs 17:30:01 Integrated Voltage Measurement... 17:30:10 Voltage Offset Correction: v_off_ini = 198.458 uV, v_off_fin = 188.100 uV 17:30:10 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -324.894 uVs 17:30:10 Integrated Voltage Measurement... 17:30:19 Voltage Offset Correction: v_off_ini = 203.196 uV, v_off_fin = 211.379 uV 17:30:19 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 295.141 uVs 17:30:22 I2Y Average and RMS of Measurements VT = -315.132 +- 7.790 uVs I2Y = -28.993 +- 0.716 uTm^2 17:30:26 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 17:30:30 Integrated Voltage Measurement... 17:30:39 Voltage Offset Correction: v_off_ini = 214.044 uV, v_off_fin = 205.868 uV 17:30:39 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -85.380 uVs 17:30:39 Integrated Voltage Measurement... 17:30:48 Voltage Offset Correction: v_off_ini = 211.352 uV, v_off_fin = 221.193 uV 17:30:48 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 54.096 uVs 17:30:48 Integrated Voltage Measurement... 17:30:57 Voltage Offset Correction: v_off_ini = 193.751 uV, v_off_fin = 205.673 uV 17:30:57 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.4999 mm dXY = -0.4999 mm VT = -21.905 uVs 17:30:57 Integrated Voltage Measurement... 17:31:06 Voltage Offset Correction: v_off_ini = 216.360 uV, v_off_fin = 229.865 uV 17:31:06 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = 69.133 uVs 17:31:06 Integrated Voltage Measurement... 17:31:15 Voltage Offset Correction: v_off_ini = 232.793 uV, v_off_fin = 224.507 uV 17:31:15 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.4999 mm dXY = -0.4999 mm VT = -70.767 uVs 17:31:15 Integrated Voltage Measurement... 17:31:24 Voltage Offset Correction: v_off_ini = 232.621 uV, v_off_fin = 232.682 uV 17:31:24 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = 74.912 uVs 17:31:28 I1X Average and RMS of Measurements VT = -62.699 +- 12.214 uVs I1X = 12.541 +- 2.443 uTm 17:31:31 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 17:31:35 Integrated Voltage Measurement... 17:31:44 Voltage Offset Correction: v_off_ini = 232.177 uV, v_off_fin = 223.427 uV 17:31:44 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -204.228 uVs 17:31:44 Integrated Voltage Measurement... 17:31:53 Voltage Offset Correction: v_off_ini = 229.795 uV, v_off_fin = 252.106 uV 17:31:53 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 192.219 uVs 17:31:53 Integrated Voltage Measurement... 17:32:02 Voltage Offset Correction: v_off_ini = 231.336 uV, v_off_fin = 228.386 uV 17:32:02 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -196.830 uVs 17:32:02 Integrated Voltage Measurement... 17:32:11 Voltage Offset Correction: v_off_ini = 236.890 uV, v_off_fin = 237.194 uV 17:32:11 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 199.307 uVs 17:32:11 Integrated Voltage Measurement... 17:32:20 Voltage Offset Correction: v_off_ini = 232.998 uV, v_off_fin = 214.407 uV 17:32:20 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -195.945 uVs 17:32:20 Integrated Voltage Measurement... 17:32:29 Voltage Offset Correction: v_off_ini = 239.419 uV, v_off_fin = 245.279 uV 17:32:29 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 192.654 uVs 17:32:33 I1Y Average and RMS of Measurements VT = -196.864 +- 1.814 uVs I1Y = -39.373 +- 0.363 uTm 17:32:35 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 17:32:38 Integrated Voltage Measurement... 17:32:47 Voltage Offset Correction: v_off_ini = 236.473 uV, v_off_fin = 223.600 uV 17:32:47 Integrated Voltage Measurement dXYus = -1.0001 mm, dXYds = 0.0000 mm dXY = -1.0001 mm VT = -87.450 uVs 17:32:47 Integrated Voltage Measurement... 17:32:56 Voltage Offset Correction: v_off_ini = 250.440 uV, v_off_fin = 253.648 uV 17:32:56 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 67.797 uVs 17:32:56 Integrated Voltage Measurement... 17:33:05 Voltage Offset Correction: v_off_ini = 255.386 uV, v_off_fin = 242.886 uV 17:33:05 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -77.383 uVs 17:33:05 Integrated Voltage Measurement... 17:33:14 Voltage Offset Correction: v_off_ini = 259.371 uV, v_off_fin = 264.023 uV 17:33:14 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 81.271 uVs 17:33:14 Integrated Voltage Measurement... 17:33:23 Voltage Offset Correction: v_off_ini = 262.699 uV, v_off_fin = 259.351 uV 17:33:23 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -69.676 uVs 17:33:23 Integrated Voltage Measurement... 17:33:32 Voltage Offset Correction: v_off_ini = 268.393 uV, v_off_fin = 261.513 uV 17:33:32 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 65.556 uVs 17:33:34 I2X Average and RMS of Measurements VT = -74.855 +- 5.166 uVs I2X = 6.887 +- 0.475 uTm^2 17:33:37 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 17:33:39 Integrated Voltage Measurement... 17:33:48 Voltage Offset Correction: v_off_ini = 253.378 uV, v_off_fin = 252.553 uV 17:33:48 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -317.936 uVs 17:33:48 Integrated Voltage Measurement... 17:33:57 Voltage Offset Correction: v_off_ini = 251.901 uV, v_off_fin = 243.366 uV 17:33:57 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 324.259 uVs 17:33:58 Integrated Voltage Measurement... 17:34:06 Voltage Offset Correction: v_off_ini = 241.537 uV, v_off_fin = 249.928 uV 17:34:06 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -310.775 uVs 17:34:07 Integrated Voltage Measurement... 17:34:15 Voltage Offset Correction: v_off_ini = 242.693 uV, v_off_fin = 249.207 uV 17:34:15 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 316.009 uVs 17:34:16 Integrated Voltage Measurement... 17:34:24 Voltage Offset Correction: v_off_ini = 252.087 uV, v_off_fin = 257.480 uV 17:34:25 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -300.379 uVs 17:34:25 Integrated Voltage Measurement... 17:34:33 Voltage Offset Correction: v_off_ini = 259.696 uV, v_off_fin = 258.666 uV 17:34:34 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 298.762 uVs 17:34:36 I2Y Average and RMS of Measurements VT = -311.353 +- 8.906 uVs I2Y = -28.645 +- 0.819 uTm^2 17:34:40 Wire Position: X0 = 0.0000 mm, Y0 = 0.5000 mm 17:34:44 Integrated Voltage Measurement... 17:34:53 Voltage Offset Correction: v_off_ini = 243.757 uV, v_off_fin = 243.188 uV 17:34:53 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -71.892 uVs 17:34:53 Integrated Voltage Measurement... 17:35:02 Voltage Offset Correction: v_off_ini = 247.739 uV, v_off_fin = 253.097 uV 17:35:02 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 69.340 uVs 17:35:02 Integrated Voltage Measurement... 17:35:11 Voltage Offset Correction: v_off_ini = 252.654 uV, v_off_fin = 248.113 uV 17:35:11 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -63.039 uVs 17:35:11 Integrated Voltage Measurement... 17:35:20 Voltage Offset Correction: v_off_ini = 235.144 uV, v_off_fin = 243.998 uV 17:35:20 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = 61.287 uVs 17:35:20 Integrated Voltage Measurement... 17:35:29 Voltage Offset Correction: v_off_ini = 247.000 uV, v_off_fin = 245.537 uV 17:35:29 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -57.900 uVs 17:35:29 Integrated Voltage Measurement... 17:35:38 Voltage Offset Correction: v_off_ini = 248.693 uV, v_off_fin = 249.661 uV 17:35:38 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 66.738 uVs 17:35:42 I1X Average and RMS of Measurements VT = -65.033 +- 3.948 uVs I1X = 13.007 +- 0.790 uTm 17:35:45 Wire Position: X0 = 0.0000 mm, Y0 = 0.5000 mm 17:35:49 Integrated Voltage Measurement... 17:35:58 Voltage Offset Correction: v_off_ini = 248.712 uV, v_off_fin = 237.719 uV 17:35:58 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -190.458 uVs 17:35:58 Integrated Voltage Measurement... 17:36:07 Voltage Offset Correction: v_off_ini = 252.903 uV, v_off_fin = 271.120 uV 17:36:07 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 204.279 uVs 17:36:07 Integrated Voltage Measurement... 17:36:16 Voltage Offset Correction: v_off_ini = 258.298 uV, v_off_fin = 243.159 uV 17:36:16 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -206.847 uVs 17:36:16 Integrated Voltage Measurement... 17:36:25 Voltage Offset Correction: v_off_ini = 255.585 uV, v_off_fin = 268.219 uV 17:36:25 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 194.423 uVs 17:36:25 Integrated Voltage Measurement... 17:36:34 Voltage Offset Correction: v_off_ini = 246.706 uV, v_off_fin = 243.323 uV 17:36:34 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4998 mm dXY = 0.4999 mm VT = -181.496 uVs 17:36:34 Integrated Voltage Measurement... 17:36:43 Voltage Offset Correction: v_off_ini = 252.567 uV, v_off_fin = 263.487 uV 17:36:43 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4998 mm dXY = -0.4999 mm VT = 180.690 uVs 17:36:47 I1Y Average and RMS of Measurements VT = -193.032 +- 8.547 uVs I1Y = -38.608 +- 1.705 uTm 17:36:50 Wire Position: X0 = 0.0000 mm, Y0 = 0.5000 mm 17:36:52 Integrated Voltage Measurement... 17:37:01 Voltage Offset Correction: v_off_ini = 234.990 uV, v_off_fin = 245.084 uV 17:37:01 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -54.185 uVs 17:37:01 Integrated Voltage Measurement... 17:37:10 Voltage Offset Correction: v_off_ini = 244.602 uV, v_off_fin = 260.806 uV 17:37:10 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 59.977 uVs 17:37:10 Integrated Voltage Measurement... 17:37:19 Voltage Offset Correction: v_off_ini = 255.099 uV, v_off_fin = 245.419 uV 17:37:19 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -74.112 uVs 17:37:19 Integrated Voltage Measurement... 17:37:28 Voltage Offset Correction: v_off_ini = 266.310 uV, v_off_fin = 262.592 uV 17:37:28 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 56.567 uVs 17:37:28 Integrated Voltage Measurement... 17:37:37 Voltage Offset Correction: v_off_ini = 263.969 uV, v_off_fin = 244.119 uV 17:37:37 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -70.114 uVs 17:37:37 Integrated Voltage Measurement... 17:37:46 Voltage Offset Correction: v_off_ini = 256.143 uV, v_off_fin = 255.496 uV 17:37:46 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 55.690 uVs 17:37:49 I2X Average and RMS of Measurements VT = -61.774 +- 3.464 uVs I2X = 5.683 +- 0.319 uTm^2 17:37:51 Wire Position: X0 = 0.0000 mm, Y0 = 0.5000 mm 17:37:54 Integrated Voltage Measurement... 17:38:03 Voltage Offset Correction: v_off_ini = 253.641 uV, v_off_fin = 242.966 uV 17:38:03 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -330.098 uVs 17:38:03 Integrated Voltage Measurement... 17:38:12 Voltage Offset Correction: v_off_ini = 250.351 uV, v_off_fin = 260.325 uV 17:38:12 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 304.437 uVs 17:38:12 Integrated Voltage Measurement... 17:38:21 Voltage Offset Correction: v_off_ini = 270.345 uV, v_off_fin = 250.466 uV 17:38:21 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = -319.007 uVs 17:38:21 Integrated Voltage Measurement... 17:38:30 Voltage Offset Correction: v_off_ini = 254.050 uV, v_off_fin = 243.555 uV 17:38:30 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 309.307 uVs 17:38:30 Integrated Voltage Measurement... 17:38:39 Voltage Offset Correction: v_off_ini = 242.439 uV, v_off_fin = 237.204 uV 17:38:39 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = -306.389 uVs 17:38:39 Integrated Voltage Measurement... 17:38:48 Voltage Offset Correction: v_off_ini = 238.036 uV, v_off_fin = 246.042 uV 17:38:48 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 300.456 uVs 17:38:50 I2Y Average and RMS of Measurements VT = -311.616 +- 5.931 uVs I2Y = -28.670 +- 0.545 uTm^2 17:38:54 Wire Position: X0 = 0.0000 mm, Y0 = 1.0000 mm 17:38:58 Integrated Voltage Measurement... 17:39:07 Voltage Offset Correction: v_off_ini = 260.443 uV, v_off_fin = 238.956 uV 17:39:07 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -69.460 uVs 17:39:07 Integrated Voltage Measurement... 17:39:16 Voltage Offset Correction: v_off_ini = 235.704 uV, v_off_fin = 243.063 uV 17:39:16 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 41.007 uVs 17:39:16 Integrated Voltage Measurement... 17:39:25 Voltage Offset Correction: v_off_ini = 234.545 uV, v_off_fin = 237.439 uV 17:39:25 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -60.701 uVs 17:39:25 Integrated Voltage Measurement... 17:39:34 Voltage Offset Correction: v_off_ini = 248.557 uV, v_off_fin = 253.187 uV 17:39:34 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 71.746 uVs 17:39:34 Integrated Voltage Measurement... 17:39:43 Voltage Offset Correction: v_off_ini = 238.562 uV, v_off_fin = 247.558 uV 17:39:43 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -74.930 uVs 17:39:43 Integrated Voltage Measurement... 17:39:52 Voltage Offset Correction: v_off_ini = 244.440 uV, v_off_fin = 254.314 uV 17:39:52 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 53.798 uVs 17:39:57 I1X Average and RMS of Measurements VT = -61.940 +- 4.803 uVs I1X = 12.388 +- 0.961 uTm 17:39:59 Wire Position: X0 = 0.0000 mm, Y0 = 1.0000 mm 17:40:03 Integrated Voltage Measurement... 17:40:12 Voltage Offset Correction: v_off_ini = 242.600 uV, v_off_fin = 223.107 uV 17:40:12 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -182.813 uVs 17:40:12 Integrated Voltage Measurement... 17:40:21 Voltage Offset Correction: v_off_ini = 244.100 uV, v_off_fin = 248.220 uV 17:40:21 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 176.138 uVs 17:40:21 Integrated Voltage Measurement... 17:40:30 Voltage Offset Correction: v_off_ini = 255.807 uV, v_off_fin = 230.096 uV 17:40:30 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -190.440 uVs 17:40:30 Integrated Voltage Measurement... 17:40:39 Voltage Offset Correction: v_off_ini = 233.236 uV, v_off_fin = 244.816 uV 17:40:39 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 189.511 uVs 17:40:39 Integrated Voltage Measurement... 17:40:48 Voltage Offset Correction: v_off_ini = 239.600 uV, v_off_fin = 234.261 uV 17:40:48 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -186.752 uVs 17:40:48 Integrated Voltage Measurement... 17:40:57 Voltage Offset Correction: v_off_ini = 259.455 uV, v_off_fin = 268.695 uV 17:40:57 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 189.659 uVs 17:41:01 I1Y Average and RMS of Measurements VT = -185.886 +- 4.590 uVs I1Y = -37.176 +- 0.918 uTm 17:41:04 Wire Position: X0 = 0.0000 mm, Y0 = 1.0000 mm 17:41:06 Integrated Voltage Measurement... 17:41:15 Voltage Offset Correction: v_off_ini = 240.236 uV, v_off_fin = 238.632 uV 17:41:15 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -55.782 uVs 17:41:15 Integrated Voltage Measurement... 17:41:24 Voltage Offset Correction: v_off_ini = 246.280 uV, v_off_fin = 254.966 uV 17:41:24 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 51.149 uVs 17:41:24 Integrated Voltage Measurement... 17:41:33 Voltage Offset Correction: v_off_ini = 251.603 uV, v_off_fin = 241.847 uV 17:41:33 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -68.951 uVs 17:41:33 Integrated Voltage Measurement... 17:41:42 Voltage Offset Correction: v_off_ini = 254.994 uV, v_off_fin = 262.032 uV 17:41:42 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 59.603 uVs 17:41:42 Integrated Voltage Measurement... 17:41:51 Voltage Offset Correction: v_off_ini = 256.614 uV, v_off_fin = 246.477 uV 17:41:51 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -63.118 uVs 17:41:51 Integrated Voltage Measurement... 17:42:00 Voltage Offset Correction: v_off_ini = 248.971 uV, v_off_fin = 261.305 uV 17:42:00 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 73.567 uVs 17:42:03 I2X Average and RMS of Measurements VT = -62.028 +- 6.278 uVs I2X = 5.707 +- 0.578 uTm^2 17:42:05 Wire Position: X0 = 0.0000 mm, Y0 = 1.0000 mm 17:42:08 Integrated Voltage Measurement... 17:42:17 Voltage Offset Correction: v_off_ini = 255.033 uV, v_off_fin = 253.212 uV 17:42:17 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -293.780 uVs 17:42:17 Integrated Voltage Measurement... 17:42:26 Voltage Offset Correction: v_off_ini = 269.402 uV, v_off_fin = 247.138 uV 17:42:26 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 293.258 uVs 17:42:26 Integrated Voltage Measurement... 17:42:35 Voltage Offset Correction: v_off_ini = 250.636 uV, v_off_fin = 250.197 uV 17:42:35 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -313.128 uVs 17:42:35 Integrated Voltage Measurement... 17:42:44 Voltage Offset Correction: v_off_ini = 254.133 uV, v_off_fin = 255.781 uV 17:42:44 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 316.949 uVs 17:42:44 Integrated Voltage Measurement... 17:42:53 Voltage Offset Correction: v_off_ini = 245.246 uV, v_off_fin = 242.500 uV 17:42:53 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -303.933 uVs 17:42:53 Integrated Voltage Measurement... 17:43:02 Voltage Offset Correction: v_off_ini = 247.469 uV, v_off_fin = 244.903 uV 17:43:02 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 312.852 uVs 17:43:04 I2Y Average and RMS of Measurements VT = -305.650 +- 8.997 uVs I2Y = -28.120 +- 0.828 uTm^2 17:43:21 Wire Position: X0 = -2.0000 mm, Y0 = 0.0000 mm 17:43:26 Integrated Voltage Measurement... 17:43:34 Voltage Offset Correction: v_off_ini = 228.632 uV, v_off_fin = 216.635 uV 17:43:34 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.4998 mm dXY = -0.4999 mm VT = -83.387 uVs 17:43:35 Integrated Voltage Measurement... 17:43:43 Voltage Offset Correction: v_off_ini = 223.348 uV, v_off_fin = 231.191 uV 17:43:43 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 87.366 uVs 17:43:44 Integrated Voltage Measurement... 17:43:52 Voltage Offset Correction: v_off_ini = 239.496 uV, v_off_fin = 226.885 uV 17:43:52 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -83.730 uVs 17:43:53 Integrated Voltage Measurement... 17:44:01 Voltage Offset Correction: v_off_ini = 235.335 uV, v_off_fin = 236.632 uV 17:44:01 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 78.420 uVs 17:44:01 Integrated Voltage Measurement... 17:44:10 Voltage Offset Correction: v_off_ini = 227.938 uV, v_off_fin = 208.470 uV 17:44:10 Integrated Voltage Measurement dXYus = -0.4998 mm, dXYds = -0.4999 mm dXY = -0.4999 mm VT = -94.022 uVs 17:44:10 Integrated Voltage Measurement... 17:44:19 Voltage Offset Correction: v_off_ini = 224.077 uV, v_off_fin = 223.083 uV 17:44:19 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = 91.047 uVs 17:44:24 I1X Average and RMS of Measurements VT = -86.329 +- 4.727 uVs I1X = 17.268 +- 0.946 uTm 17:44:26 Wire Position: X0 = -2.0000 mm, Y0 = 0.0000 mm 17:44:30 Integrated Voltage Measurement... 17:44:39 Voltage Offset Correction: v_off_ini = 227.973 uV, v_off_fin = 222.386 uV 17:44:39 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -189.238 uVs 17:44:39 Integrated Voltage Measurement... 17:44:48 Voltage Offset Correction: v_off_ini = 226.839 uV, v_off_fin = 236.802 uV 17:44:48 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 189.104 uVs 17:44:48 Integrated Voltage Measurement... 17:44:57 Voltage Offset Correction: v_off_ini = 228.773 uV, v_off_fin = 222.182 uV 17:44:57 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -193.465 uVs 17:44:57 Integrated Voltage Measurement... 17:45:06 Voltage Offset Correction: v_off_ini = 239.341 uV, v_off_fin = 251.501 uV 17:45:06 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 180.662 uVs 17:45:06 Integrated Voltage Measurement... 17:45:15 Voltage Offset Correction: v_off_ini = 239.795 uV, v_off_fin = 229.011 uV 17:45:15 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -191.847 uVs 17:45:15 Integrated Voltage Measurement... 17:45:24 Voltage Offset Correction: v_off_ini = 231.516 uV, v_off_fin = 232.404 uV 17:45:24 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 175.981 uVs 17:45:28 I1Y Average and RMS of Measurements VT = -186.716 +- 2.160 uVs I1Y = -37.343 +- 0.432 uTm 17:45:33 Wire Position: X0 = -2.0000 mm, Y0 = 0.0000 mm 17:45:36 Integrated Voltage Measurement... 17:45:45 Voltage Offset Correction: v_off_ini = 238.083 uV, v_off_fin = 229.328 uV 17:45:45 Integrated Voltage Measurement dXYus = -1.0001 mm, dXYds = 0.0000 mm dXY = -1.0001 mm VT = -116.738 uVs 17:45:45 Integrated Voltage Measurement... 17:45:54 Voltage Offset Correction: v_off_ini = 239.847 uV, v_off_fin = 242.792 uV 17:45:54 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 122.884 uVs 17:45:54 Integrated Voltage Measurement... 17:46:03 Voltage Offset Correction: v_off_ini = 227.541 uV, v_off_fin = 220.664 uV 17:46:03 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -100.237 uVs 17:46:03 Integrated Voltage Measurement... 17:46:12 Voltage Offset Correction: v_off_ini = 247.020 uV, v_off_fin = 252.371 uV 17:46:12 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 85.033 uVs 17:46:12 Integrated Voltage Measurement... 17:46:21 Voltage Offset Correction: v_off_ini = 230.807 uV, v_off_fin = 221.123 uV 17:46:21 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -123.502 uVs 17:46:21 Integrated Voltage Measurement... 17:46:30 Voltage Offset Correction: v_off_ini = 236.556 uV, v_off_fin = 232.778 uV 17:46:30 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 111.182 uVs 17:46:32 I2X Average and RMS of Measurements VT = -109.929 +- 12.270 uVs I2X = 10.113 +- 1.129 uTm^2 17:46:38 Wire Position: X0 = -2.0000 mm, Y0 = 0.0000 mm 17:46:40 Integrated Voltage Measurement... 17:46:49 Voltage Offset Correction: v_off_ini = 233.004 uV, v_off_fin = 231.024 uV 17:46:49 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -297.614 uVs 17:46:49 Integrated Voltage Measurement... 17:46:58 Voltage Offset Correction: v_off_ini = 251.185 uV, v_off_fin = 245.440 uV 17:46:58 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 306.435 uVs 17:46:58 Integrated Voltage Measurement... 17:47:07 Voltage Offset Correction: v_off_ini = 239.181 uV, v_off_fin = 240.255 uV 17:47:07 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -295.201 uVs 17:47:07 Integrated Voltage Measurement... 17:47:16 Voltage Offset Correction: v_off_ini = 238.024 uV, v_off_fin = 244.759 uV 17:47:16 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 310.343 uVs 17:47:16 Integrated Voltage Measurement... 17:47:25 Voltage Offset Correction: v_off_ini = 242.584 uV, v_off_fin = 242.358 uV 17:47:25 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -297.318 uVs 17:47:25 Integrated Voltage Measurement... 17:47:34 Voltage Offset Correction: v_off_ini = 243.373 uV, v_off_fin = 253.145 uV 17:47:34 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 314.266 uVs 17:47:37 I2Y Average and RMS of Measurements VT = -303.530 +- 1.629 uVs I2Y = -27.925 +- 0.150 uTm^2 17:47:41 Wire Position: X0 = -1.0000 mm, Y0 = 0.0000 mm 17:47:45 Integrated Voltage Measurement... 17:47:54 Voltage Offset Correction: v_off_ini = 260.128 uV, v_off_fin = 245.866 uV 17:47:54 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -93.166 uVs 17:47:54 Integrated Voltage Measurement... 17:48:03 Voltage Offset Correction: v_off_ini = 247.259 uV, v_off_fin = 252.628 uV 17:48:03 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 80.261 uVs 17:48:03 Integrated Voltage Measurement... 17:48:12 Voltage Offset Correction: v_off_ini = 259.645 uV, v_off_fin = 254.444 uV 17:48:12 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -65.585 uVs 17:48:12 Integrated Voltage Measurement... 17:48:21 Voltage Offset Correction: v_off_ini = 262.562 uV, v_off_fin = 264.056 uV 17:48:21 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 87.946 uVs 17:48:21 Integrated Voltage Measurement... 17:48:30 Voltage Offset Correction: v_off_ini = 250.914 uV, v_off_fin = 254.537 uV 17:48:30 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -93.953 uVs 17:48:30 Integrated Voltage Measurement... 17:48:39 Voltage Offset Correction: v_off_ini = 260.962 uV, v_off_fin = 262.945 uV 17:48:39 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 80.403 uVs 17:48:44 I1X Average and RMS of Measurements VT = -83.553 +- 4.803 uVs I1X = 16.711 +- 0.961 uTm 17:48:46 Wire Position: X0 = -1.0000 mm, Y0 = 0.0000 mm 17:48:50 Integrated Voltage Measurement... 17:48:59 Voltage Offset Correction: v_off_ini = 241.524 uV, v_off_fin = 246.445 uV 17:48:59 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -181.011 uVs 17:48:59 Integrated Voltage Measurement... 17:49:08 Voltage Offset Correction: v_off_ini = 257.153 uV, v_off_fin = 267.118 uV 17:49:08 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 196.508 uVs 17:49:08 Integrated Voltage Measurement... 17:49:17 Voltage Offset Correction: v_off_ini = 264.910 uV, v_off_fin = 250.687 uV 17:49:17 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -195.226 uVs 17:49:17 Integrated Voltage Measurement... 17:49:26 Voltage Offset Correction: v_off_ini = 260.133 uV, v_off_fin = 277.997 uV 17:49:26 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 207.665 uVs 17:49:26 Integrated Voltage Measurement... 17:49:35 Voltage Offset Correction: v_off_ini = 280.636 uV, v_off_fin = 256.938 uV 17:49:35 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -187.143 uVs 17:49:35 Integrated Voltage Measurement... 17:49:44 Voltage Offset Correction: v_off_ini = 271.601 uV, v_off_fin = 280.407 uV 17:49:44 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 202.382 uVs 17:49:48 I1Y Average and RMS of Measurements VT = -194.989 +- 5.182 uVs I1Y = -38.998 +- 1.036 uTm 17:49:53 Wire Position: X0 = -1.0000 mm, Y0 = 0.0000 mm 17:49:56 Integrated Voltage Measurement... 17:50:05 Voltage Offset Correction: v_off_ini = 265.734 uV, v_off_fin = 246.803 uV 17:50:05 Integrated Voltage Measurement dXYus = -1.0001 mm, dXYds = 0.0000 mm dXY = -1.0001 mm VT = -105.414 uVs 17:50:05 Integrated Voltage Measurement... 17:50:14 Voltage Offset Correction: v_off_ini = 267.274 uV, v_off_fin = 271.610 uV 17:50:14 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 98.714 uVs 17:50:14 Integrated Voltage Measurement... 17:50:23 Voltage Offset Correction: v_off_ini = 256.501 uV, v_off_fin = 253.563 uV 17:50:23 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -89.723 uVs 17:50:23 Integrated Voltage Measurement... 17:50:32 Voltage Offset Correction: v_off_ini = 264.003 uV, v_off_fin = 283.842 uV 17:50:32 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 106.188 uVs 17:50:32 Integrated Voltage Measurement... 17:50:41 Voltage Offset Correction: v_off_ini = 261.767 uV, v_off_fin = 254.168 uV 17:50:41 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -92.785 uVs 17:50:41 Integrated Voltage Measurement... 17:50:50 Voltage Offset Correction: v_off_ini = 259.472 uV, v_off_fin = 268.741 uV 17:50:50 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 105.899 uVs 17:50:52 I2X Average and RMS of Measurements VT = -99.787 +- 1.707 uVs I2X = 9.180 +- 0.157 uTm^2 17:50:57 Wire Position: X0 = -1.0000 mm, Y0 = 0.0000 mm 17:51:00 Integrated Voltage Measurement... 17:51:09 Voltage Offset Correction: v_off_ini = 265.226 uV, v_off_fin = 264.410 uV 17:51:09 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -300.967 uVs 17:51:09 Integrated Voltage Measurement... 17:51:18 Voltage Offset Correction: v_off_ini = 273.935 uV, v_off_fin = 256.607 uV 17:51:18 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 293.745 uVs 17:51:18 Integrated Voltage Measurement... 17:51:27 Voltage Offset Correction: v_off_ini = 265.699 uV, v_off_fin = 268.806 uV 17:51:27 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -308.466 uVs 17:51:27 Integrated Voltage Measurement... 17:51:36 Voltage Offset Correction: v_off_ini = 273.284 uV, v_off_fin = 269.493 uV 17:51:36 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 293.225 uVs 17:51:36 Integrated Voltage Measurement... 17:51:45 Voltage Offset Correction: v_off_ini = 269.257 uV, v_off_fin = 269.562 uV 17:51:45 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -297.642 uVs 17:51:45 Integrated Voltage Measurement... 17:51:54 Voltage Offset Correction: v_off_ini = 274.743 uV, v_off_fin = 265.738 uV 17:51:54 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 292.763 uVs 17:51:56 I2Y Average and RMS of Measurements VT = -297.801 +- 2.325 uVs I2Y = -27.398 +- 0.214 uTm^2 17:52:01 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 17:52:05 Integrated Voltage Measurement... 17:52:14 Voltage Offset Correction: v_off_ini = 271.330 uV, v_off_fin = 274.266 uV 17:52:14 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -87.988 uVs 17:52:14 Integrated Voltage Measurement... 17:52:23 Voltage Offset Correction: v_off_ini = 265.742 uV, v_off_fin = 274.900 uV 17:52:23 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 73.958 uVs 17:52:23 Integrated Voltage Measurement... 17:52:32 Voltage Offset Correction: v_off_ini = 247.514 uV, v_off_fin = 261.679 uV 17:52:32 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -61.964 uVs 17:52:32 Integrated Voltage Measurement... 17:52:41 Voltage Offset Correction: v_off_ini = 257.010 uV, v_off_fin = 259.113 uV 17:52:41 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 96.293 uVs 17:52:41 Integrated Voltage Measurement... 17:52:50 Voltage Offset Correction: v_off_ini = 267.579 uV, v_off_fin = 265.309 uV 17:52:50 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -87.840 uVs 17:52:50 Integrated Voltage Measurement... 17:52:59 Voltage Offset Correction: v_off_ini = 262.012 uV, v_off_fin = 263.659 uV 17:52:59 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 83.504 uVs 17:53:03 I1X Average and RMS of Measurements VT = -81.924 +- 2.755 uVs I1X = 16.385 +- 0.551 uTm 17:53:06 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 17:53:10 Integrated Voltage Measurement... 17:53:19 Voltage Offset Correction: v_off_ini = 256.268 uV, v_off_fin = 256.999 uV 17:53:19 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -186.483 uVs 17:53:19 Integrated Voltage Measurement... 17:53:28 Voltage Offset Correction: v_off_ini = 259.979 uV, v_off_fin = 270.885 uV 17:53:28 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 171.246 uVs 17:53:28 Integrated Voltage Measurement... 17:53:37 Voltage Offset Correction: v_off_ini = 257.216 uV, v_off_fin = 256.805 uV 17:53:37 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -190.506 uVs 17:53:37 Integrated Voltage Measurement... 17:53:46 Voltage Offset Correction: v_off_ini = 249.699 uV, v_off_fin = 270.783 uV 17:53:46 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 200.162 uVs 17:53:46 Integrated Voltage Measurement... 17:53:55 Voltage Offset Correction: v_off_ini = 257.461 uV, v_off_fin = 246.215 uV 17:53:55 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -186.467 uVs 17:53:55 Integrated Voltage Measurement... 17:54:04 Voltage Offset Correction: v_off_ini = 262.732 uV, v_off_fin = 276.661 uV 17:54:04 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 184.065 uVs 17:54:08 I1Y Average and RMS of Measurements VT = -186.488 +- 6.779 uVs I1Y = -37.298 +- 1.356 uTm 17:54:13 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 17:54:16 Integrated Voltage Measurement... 17:54:25 Voltage Offset Correction: v_off_ini = 257.073 uV, v_off_fin = 253.664 uV 17:54:25 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -84.002 uVs 17:54:25 Integrated Voltage Measurement... 17:54:34 Voltage Offset Correction: v_off_ini = 251.273 uV, v_off_fin = 265.087 uV 17:54:34 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 96.304 uVs 17:54:34 Integrated Voltage Measurement... 17:54:43 Voltage Offset Correction: v_off_ini = 260.238 uV, v_off_fin = 249.088 uV 17:54:43 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -91.310 uVs 17:54:43 Integrated Voltage Measurement... 17:54:52 Voltage Offset Correction: v_off_ini = 265.839 uV, v_off_fin = 276.358 uV 17:54:52 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 88.169 uVs 17:54:52 Integrated Voltage Measurement... 17:55:01 Voltage Offset Correction: v_off_ini = 262.873 uV, v_off_fin = 250.155 uV 17:55:01 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -80.030 uVs 17:55:01 Integrated Voltage Measurement... 17:55:10 Voltage Offset Correction: v_off_ini = 272.815 uV, v_off_fin = 285.396 uV 17:55:10 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 84.212 uVs 17:55:12 I2X Average and RMS of Measurements VT = -87.338 +- 3.693 uVs I2X = 8.035 +- 0.340 uTm^2 17:55:17 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 17:55:20 Integrated Voltage Measurement... 17:55:29 Voltage Offset Correction: v_off_ini = 269.782 uV, v_off_fin = 275.829 uV 17:55:29 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -287.810 uVs 17:55:29 Integrated Voltage Measurement... 17:55:38 Voltage Offset Correction: v_off_ini = 276.450 uV, v_off_fin = 283.894 uV 17:55:38 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 301.459 uVs 17:55:38 Integrated Voltage Measurement... 17:55:47 Voltage Offset Correction: v_off_ini = 277.953 uV, v_off_fin = 265.276 uV 17:55:47 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -303.915 uVs 17:55:47 Integrated Voltage Measurement... 17:55:56 Voltage Offset Correction: v_off_ini = 264.683 uV, v_off_fin = 271.530 uV 17:55:56 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 304.684 uVs 17:55:56 Integrated Voltage Measurement... 17:56:05 Voltage Offset Correction: v_off_ini = 253.221 uV, v_off_fin = 261.836 uV 17:56:05 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -303.679 uVs 17:56:05 Integrated Voltage Measurement... 17:56:14 Voltage Offset Correction: v_off_ini = 264.667 uV, v_off_fin = 270.391 uV 17:56:14 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 305.237 uVs 17:56:16 I2Y Average and RMS of Measurements VT = -301.131 +- 4.594 uVs I2Y = -27.704 +- 0.423 uTm^2 17:56:21 Wire Position: X0 = 1.0000 mm, Y0 = 0.0000 mm 17:56:25 Integrated Voltage Measurement... 17:56:34 Voltage Offset Correction: v_off_ini = 252.806 uV, v_off_fin = 241.533 uV 17:56:34 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -74.628 uVs 17:56:34 Integrated Voltage Measurement... 17:56:43 Voltage Offset Correction: v_off_ini = 245.898 uV, v_off_fin = 258.087 uV 17:56:43 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 72.530 uVs 17:56:43 Integrated Voltage Measurement... 17:56:52 Voltage Offset Correction: v_off_ini = 243.537 uV, v_off_fin = 254.515 uV 17:56:52 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -38.004 uVs 17:56:52 Integrated Voltage Measurement... 17:57:01 Voltage Offset Correction: v_off_ini = 277.630 uV, v_off_fin = 273.385 uV 17:57:01 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 68.101 uVs 17:57:01 Integrated Voltage Measurement... 17:57:10 Voltage Offset Correction: v_off_ini = 269.034 uV, v_off_fin = 251.730 uV 17:57:10 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.4999 mm dXY = -0.4999 mm VT = -89.850 uVs 17:57:10 Integrated Voltage Measurement... 17:57:19 Voltage Offset Correction: v_off_ini = 265.371 uV, v_off_fin = 263.071 uV 17:57:19 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = 61.207 uVs 17:57:23 I1X Average and RMS of Measurements VT = -67.387 +- 10.167 uVs I1X = 13.478 +- 2.034 uTm 17:57:26 Wire Position: X0 = 1.0000 mm, Y0 = 0.0000 mm 17:57:30 Integrated Voltage Measurement... 17:57:39 Voltage Offset Correction: v_off_ini = 247.349 uV, v_off_fin = 242.332 uV 17:57:39 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -183.976 uVs 17:57:39 Integrated Voltage Measurement... 17:57:48 Voltage Offset Correction: v_off_ini = 223.362 uV, v_off_fin = 243.263 uV 17:57:48 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 206.147 uVs 17:57:48 Integrated Voltage Measurement... 17:57:57 Voltage Offset Correction: v_off_ini = 231.326 uV, v_off_fin = 239.462 uV 17:57:57 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -207.202 uVs 17:57:57 Integrated Voltage Measurement... 17:58:06 Voltage Offset Correction: v_off_ini = 243.859 uV, v_off_fin = 252.982 uV 17:58:06 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 183.790 uVs 17:58:06 Integrated Voltage Measurement... 17:58:15 Voltage Offset Correction: v_off_ini = 232.719 uV, v_off_fin = 236.183 uV 17:58:15 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -191.333 uVs 17:58:15 Integrated Voltage Measurement... 17:58:24 Voltage Offset Correction: v_off_ini = 229.585 uV, v_off_fin = 244.877 uV 17:58:24 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 205.973 uVs 17:58:28 I1Y Average and RMS of Measurements VT = -196.403 +- 1.600 uVs I1Y = -39.281 +- 0.320 uTm 17:58:33 Wire Position: X0 = 1.0000 mm, Y0 = 0.0000 mm 17:58:36 Integrated Voltage Measurement... 17:58:45 Voltage Offset Correction: v_off_ini = 227.838 uV, v_off_fin = 229.468 uV 17:58:45 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -68.985 uVs 17:58:45 Integrated Voltage Measurement... 17:58:54 Voltage Offset Correction: v_off_ini = 227.012 uV, v_off_fin = 235.324 uV 17:58:54 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 95.251 uVs 17:58:54 Integrated Voltage Measurement... 17:59:03 Voltage Offset Correction: v_off_ini = 244.471 uV, v_off_fin = 229.990 uV 17:59:03 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -98.855 uVs 17:59:03 Integrated Voltage Measurement... 17:59:12 Voltage Offset Correction: v_off_ini = 239.720 uV, v_off_fin = 243.058 uV 17:59:12 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 89.855 uVs 17:59:12 Integrated Voltage Measurement... 17:59:21 Voltage Offset Correction: v_off_ini = 240.911 uV, v_off_fin = 228.103 uV 17:59:21 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -87.344 uVs 17:59:21 Integrated Voltage Measurement... 17:59:30 Voltage Offset Correction: v_off_ini = 228.702 uV, v_off_fin = 241.022 uV 17:59:30 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 92.430 uVs 17:59:32 I2X Average and RMS of Measurements VT = -88.787 +- 5.056 uVs I2X = 8.168 +- 0.465 uTm^2 17:59:37 Wire Position: X0 = 1.0000 mm, Y0 = 0.0000 mm 17:59:40 Integrated Voltage Measurement... 17:59:49 Voltage Offset Correction: v_off_ini = 228.413 uV, v_off_fin = 233.022 uV 17:59:49 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -290.066 uVs 17:59:49 Integrated Voltage Measurement... 17:59:58 Voltage Offset Correction: v_off_ini = 234.294 uV, v_off_fin = 240.089 uV 17:59:58 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 307.540 uVs 17:59:58 Integrated Voltage Measurement... 18:00:07 Voltage Offset Correction: v_off_ini = 242.319 uV, v_off_fin = 232.114 uV 18:00:07 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -302.055 uVs 18:00:07 Integrated Voltage Measurement... 18:00:16 Voltage Offset Correction: v_off_ini = 231.296 uV, v_off_fin = 235.894 uV 18:00:16 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 302.277 uVs 18:00:16 Integrated Voltage Measurement... 18:00:25 Voltage Offset Correction: v_off_ini = 225.880 uV, v_off_fin = 224.036 uV 18:00:25 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -286.863 uVs 18:00:25 Integrated Voltage Measurement... 18:00:34 Voltage Offset Correction: v_off_ini = 224.188 uV, v_off_fin = 219.268 uV 18:00:34 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 296.788 uVs 18:00:36 I2Y Average and RMS of Measurements VT = -297.598 +- 4.306 uVs I2Y = -27.379 +- 0.396 uTm^2 18:00:41 Wire Position: X0 = 2.0000 mm, Y0 = 0.0000 mm 18:00:45 Integrated Voltage Measurement... 18:00:54 Voltage Offset Correction: v_off_ini = 202.853 uV, v_off_fin = 206.867 uV 18:00:54 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -54.308 uVs 18:00:54 Integrated Voltage Measurement... 18:01:03 Voltage Offset Correction: v_off_ini = 224.664 uV, v_off_fin = 217.113 uV 18:01:03 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 45.794 uVs 18:01:03 Integrated Voltage Measurement... 18:01:12 Voltage Offset Correction: v_off_ini = 212.145 uV, v_off_fin = 215.906 uV 18:01:12 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -69.915 uVs 18:01:12 Integrated Voltage Measurement... 18:01:21 Voltage Offset Correction: v_off_ini = 207.231 uV, v_off_fin = 216.220 uV 18:01:21 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 80.675 uVs 18:01:21 Integrated Voltage Measurement... 18:01:30 Voltage Offset Correction: v_off_ini = 216.549 uV, v_off_fin = 214.635 uV 18:01:30 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -71.990 uVs 18:01:30 Integrated Voltage Measurement... 18:01:39 Voltage Offset Correction: v_off_ini = 214.342 uV, v_off_fin = 210.436 uV 18:01:39 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 76.531 uVs 18:01:43 I1X Average and RMS of Measurements VT = -66.536 +- 11.664 uVs I1X = 13.307 +- 2.333 uTm 18:01:46 Wire Position: X0 = 2.0000 mm, Y0 = 0.0000 mm 18:01:50 Integrated Voltage Measurement... 18:01:59 Voltage Offset Correction: v_off_ini = 194.543 uV, v_off_fin = 187.996 uV 18:01:59 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -178.117 uVs 18:01:59 Integrated Voltage Measurement... 18:02:08 Voltage Offset Correction: v_off_ini = 210.420 uV, v_off_fin = 214.372 uV 18:02:08 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 183.838 uVs 18:02:08 Integrated Voltage Measurement... 18:02:17 Voltage Offset Correction: v_off_ini = 206.709 uV, v_off_fin = 204.743 uV 18:02:17 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -191.578 uVs 18:02:17 Integrated Voltage Measurement... 18:02:26 Voltage Offset Correction: v_off_ini = 187.653 uV, v_off_fin = 219.292 uV 18:02:26 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 197.487 uVs 18:02:26 Integrated Voltage Measurement... 18:02:35 Voltage Offset Correction: v_off_ini = 198.774 uV, v_off_fin = 192.798 uV 18:02:35 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -182.054 uVs 18:02:35 Integrated Voltage Measurement... 18:02:44 Voltage Offset Correction: v_off_ini = 216.851 uV, v_off_fin = 221.854 uV 18:02:44 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 183.336 uVs 18:02:48 I1Y Average and RMS of Measurements VT = -186.068 +- 6.026 uVs I1Y = -37.214 +- 1.205 uTm 18:02:53 Wire Position: X0 = 2.0000 mm, Y0 = 0.0000 mm 18:02:56 Integrated Voltage Measurement... 18:03:05 Voltage Offset Correction: v_off_ini = 207.453 uV, v_off_fin = 196.387 uV 18:03:05 Integrated Voltage Measurement dXYus = -1.0001 mm, dXYds = 0.0000 mm dXY = -1.0001 mm VT = -73.912 uVs 18:03:05 Integrated Voltage Measurement... 18:03:14 Voltage Offset Correction: v_off_ini = 214.310 uV, v_off_fin = 219.414 uV 18:03:14 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 75.279 uVs 18:03:14 Integrated Voltage Measurement... 18:03:23 Voltage Offset Correction: v_off_ini = 224.584 uV, v_off_fin = 219.469 uV 18:03:23 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -83.403 uVs 18:03:23 Integrated Voltage Measurement... 18:03:32 Voltage Offset Correction: v_off_ini = 216.552 uV, v_off_fin = 232.696 uV 18:03:32 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 89.878 uVs 18:03:32 Integrated Voltage Measurement... 18:03:41 Voltage Offset Correction: v_off_ini = 222.429 uV, v_off_fin = 210.931 uV 18:03:41 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -70.615 uVs 18:03:41 Integrated Voltage Measurement... 18:03:50 Voltage Offset Correction: v_off_ini = 209.870 uV, v_off_fin = 229.534 uV 18:03:50 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 88.134 uVs 18:03:52 I2X Average and RMS of Measurements VT = -80.203 +- 4.952 uVs I2X = 7.379 +- 0.456 uTm^2 18:03:57 Wire Position: X0 = 2.0000 mm, Y0 = 0.0000 mm 18:04:00 Integrated Voltage Measurement... 18:04:09 Voltage Offset Correction: v_off_ini = 223.107 uV, v_off_fin = 226.688 uV 18:04:09 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -293.487 uVs 18:04:09 Integrated Voltage Measurement... 18:04:18 Voltage Offset Correction: v_off_ini = 220.054 uV, v_off_fin = 225.510 uV 18:04:18 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 291.474 uVs 18:04:18 Integrated Voltage Measurement... 18:04:27 Voltage Offset Correction: v_off_ini = 236.233 uV, v_off_fin = 230.237 uV 18:04:27 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -264.341 uVs 18:04:27 Integrated Voltage Measurement... 18:04:36 Voltage Offset Correction: v_off_ini = 217.583 uV, v_off_fin = 223.057 uV 18:04:36 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 309.545 uVs 18:04:36 Integrated Voltage Measurement... 18:04:45 Voltage Offset Correction: v_off_ini = 222.438 uV, v_off_fin = 219.450 uV 18:04:45 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -290.620 uVs 18:04:45 Integrated Voltage Measurement... 18:04:54 Voltage Offset Correction: v_off_ini = 222.216 uV, v_off_fin = 214.039 uV 18:04:54 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 280.663 uVs 18:04:56 I2Y Average and RMS of Measurements VT = -288.355 +- 2.965 uVs I2Y = -26.529 +- 0.273 uTm^2 18:05:04 Wire Position: X0 = 0.0000 mm, Y0 = -1.0000 mm 18:05:08 Integrated Voltage Measurement... 18:05:17 Voltage Offset Correction: v_off_ini = 227.819 uV, v_off_fin = 211.148 uV 18:05:17 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -64.754 uVs 18:05:17 Integrated Voltage Measurement... 18:05:26 Voltage Offset Correction: v_off_ini = 217.621 uV, v_off_fin = 215.807 uV 18:05:26 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 87.721 uVs 18:05:26 Integrated Voltage Measurement... 18:05:35 Voltage Offset Correction: v_off_ini = 212.560 uV, v_off_fin = 207.951 uV 18:05:35 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -86.035 uVs 18:05:35 Integrated Voltage Measurement... 18:05:44 Voltage Offset Correction: v_off_ini = 220.220 uV, v_off_fin = 216.696 uV 18:05:44 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.4999 mm dXY = 0.4999 mm VT = 51.625 uVs 18:05:44 Integrated Voltage Measurement... 18:05:53 Voltage Offset Correction: v_off_ini = 206.621 uV, v_off_fin = 214.794 uV 18:05:53 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -62.793 uVs 18:05:53 Integrated Voltage Measurement... 18:06:02 Voltage Offset Correction: v_off_ini = 214.797 uV, v_off_fin = 222.230 uV 18:06:02 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 69.316 uVs 18:06:06 I1X Average and RMS of Measurements VT = -70.374 +- 4.298 uVs I1X = 14.076 +- 0.859 uTm 18:06:09 Wire Position: X0 = 0.0000 mm, Y0 = -1.0000 mm 18:06:13 Integrated Voltage Measurement... 18:06:22 Voltage Offset Correction: v_off_ini = 214.129 uV, v_off_fin = 210.807 uV 18:06:22 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -197.109 uVs 18:06:22 Integrated Voltage Measurement... 18:06:31 Voltage Offset Correction: v_off_ini = 216.324 uV, v_off_fin = 227.751 uV 18:06:31 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 199.165 uVs 18:06:31 Integrated Voltage Measurement... 18:06:40 Voltage Offset Correction: v_off_ini = 195.222 uV, v_off_fin = 200.254 uV 18:06:40 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -193.262 uVs 18:06:40 Integrated Voltage Measurement... 18:06:49 Voltage Offset Correction: v_off_ini = 205.741 uV, v_off_fin = 225.252 uV 18:06:49 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 183.666 uVs 18:06:49 Integrated Voltage Measurement... 18:06:58 Voltage Offset Correction: v_off_ini = 220.622 uV, v_off_fin = 210.116 uV 18:06:58 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -193.203 uVs 18:06:58 Integrated Voltage Measurement... 18:07:07 Voltage Offset Correction: v_off_ini = 216.465 uV, v_off_fin = 225.356 uV 18:07:07 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 184.523 uVs 18:07:11 I1Y Average and RMS of Measurements VT = -191.821 +- 4.469 uVs I1Y = -38.364 +- 0.894 uTm 18:07:13 Wire Position: X0 = 0.0000 mm, Y0 = -1.0000 mm 18:07:16 Integrated Voltage Measurement... 18:07:25 Voltage Offset Correction: v_off_ini = 215.312 uV, v_off_fin = 211.771 uV 18:07:25 Integrated Voltage Measurement dXYus = -1.0001 mm, dXYds = 0.0000 mm dXY = -1.0001 mm VT = -80.521 uVs 18:07:25 Integrated Voltage Measurement... 18:07:34 Voltage Offset Correction: v_off_ini = 213.902 uV, v_off_fin = 219.993 uV 18:07:34 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 72.138 uVs 18:07:34 Integrated Voltage Measurement... 18:07:43 Voltage Offset Correction: v_off_ini = 218.583 uV, v_off_fin = 200.610 uV 18:07:43 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -80.411 uVs 18:07:43 Integrated Voltage Measurement... 18:07:52 Voltage Offset Correction: v_off_ini = 216.006 uV, v_off_fin = 227.899 uV 18:07:52 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 66.441 uVs 18:07:52 Integrated Voltage Measurement... 18:08:01 Voltage Offset Correction: v_off_ini = 227.479 uV, v_off_fin = 209.497 uV 18:08:01 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -100.616 uVs 18:08:01 Integrated Voltage Measurement... 18:08:10 Voltage Offset Correction: v_off_ini = 213.991 uV, v_off_fin = 211.279 uV 18:08:10 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 75.570 uVs 18:08:12 I2X Average and RMS of Measurements VT = -79.283 +- 6.341 uVs I2X = 7.294 +- 0.584 uTm^2 18:08:15 Wire Position: X0 = 0.0000 mm, Y0 = -1.0000 mm 18:08:18 Integrated Voltage Measurement... 18:08:27 Voltage Offset Correction: v_off_ini = 214.887 uV, v_off_fin = 221.210 uV 18:08:27 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -292.103 uVs 18:08:27 Integrated Voltage Measurement... 18:08:36 Voltage Offset Correction: v_off_ini = 212.034 uV, v_off_fin = 205.952 uV 18:08:36 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 318.586 uVs 18:08:36 Integrated Voltage Measurement... 18:08:45 Voltage Offset Correction: v_off_ini = 218.346 uV, v_off_fin = 213.870 uV 18:08:45 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -316.925 uVs 18:08:45 Integrated Voltage Measurement... 18:08:54 Voltage Offset Correction: v_off_ini = 208.171 uV, v_off_fin = 206.204 uV 18:08:54 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 315.974 uVs 18:08:54 Integrated Voltage Measurement... 18:09:03 Voltage Offset Correction: v_off_ini = 208.861 uV, v_off_fin = 216.803 uV 18:09:03 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -305.570 uVs 18:09:03 Integrated Voltage Measurement... 18:09:12 Voltage Offset Correction: v_off_ini = 216.251 uV, v_off_fin = 216.154 uV 18:09:12 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 311.926 uVs 18:09:14 I2Y Average and RMS of Measurements VT = -310.181 +- 4.645 uVs I2Y = -28.537 +- 0.427 uTm^2 18:09:18 Wire Position: X0 = 0.0000 mm, Y0 = -0.5000 mm 18:09:22 Integrated Voltage Measurement... 18:09:31 Voltage Offset Correction: v_off_ini = 214.674 uV, v_off_fin = 214.708 uV 18:09:31 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -88.703 uVs 18:09:31 Integrated Voltage Measurement... 18:09:40 Voltage Offset Correction: v_off_ini = 220.871 uV, v_off_fin = 212.587 uV 18:09:40 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 79.899 uVs 18:09:40 Integrated Voltage Measurement... 18:09:49 Voltage Offset Correction: v_off_ini = 219.505 uV, v_off_fin = 222.969 uV 18:09:49 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -95.444 uVs 18:09:49 Integrated Voltage Measurement... 18:09:58 Voltage Offset Correction: v_off_ini = 222.755 uV, v_off_fin = 223.310 uV 18:09:58 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 50.802 uVs 18:09:58 Integrated Voltage Measurement... 18:10:07 Voltage Offset Correction: v_off_ini = 224.290 uV, v_off_fin = 215.159 uV 18:10:07 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -83.042 uVs 18:10:07 Integrated Voltage Measurement... 18:10:16 Voltage Offset Correction: v_off_ini = 229.446 uV, v_off_fin = 222.494 uV 18:10:16 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 78.838 uVs 18:10:20 I1X Average and RMS of Measurements VT = -79.455 +- 4.683 uVs I1X = 15.891 +- 0.937 uTm 18:10:23 Wire Position: X0 = 0.0000 mm, Y0 = -0.5000 mm 18:10:27 Integrated Voltage Measurement... 18:10:36 Voltage Offset Correction: v_off_ini = 233.648 uV, v_off_fin = 228.758 uV 18:10:36 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -200.839 uVs 18:10:36 Integrated Voltage Measurement... 18:10:45 Voltage Offset Correction: v_off_ini = 238.029 uV, v_off_fin = 235.742 uV 18:10:45 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 193.850 uVs 18:10:45 Integrated Voltage Measurement... 18:10:54 Voltage Offset Correction: v_off_ini = 227.672 uV, v_off_fin = 224.786 uV 18:10:54 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -197.989 uVs 18:10:54 Integrated Voltage Measurement... 18:11:03 Voltage Offset Correction: v_off_ini = 232.027 uV, v_off_fin = 242.987 uV 18:11:03 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 186.934 uVs 18:11:03 Integrated Voltage Measurement... 18:11:12 Voltage Offset Correction: v_off_ini = 231.569 uV, v_off_fin = 217.146 uV 18:11:12 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -180.183 uVs 18:11:12 Integrated Voltage Measurement... 18:11:21 Voltage Offset Correction: v_off_ini = 222.174 uV, v_off_fin = 241.969 uV 18:11:21 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 196.024 uVs 18:11:25 I1Y Average and RMS of Measurements VT = -192.637 +- 3.775 uVs I1Y = -38.527 +- 0.755 uTm 18:11:28 Wire Position: X0 = 0.0000 mm, Y0 = -0.5000 mm 18:11:30 Integrated Voltage Measurement... 18:11:39 Voltage Offset Correction: v_off_ini = 235.678 uV, v_off_fin = 222.792 uV 18:11:39 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -92.810 uVs 18:11:39 Integrated Voltage Measurement... 18:11:48 Voltage Offset Correction: v_off_ini = 223.580 uV, v_off_fin = 230.436 uV 18:11:48 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 99.676 uVs 18:11:48 Integrated Voltage Measurement... 18:11:57 Voltage Offset Correction: v_off_ini = 235.332 uV, v_off_fin = 232.428 uV 18:11:57 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -97.202 uVs 18:11:57 Integrated Voltage Measurement... 18:12:06 Voltage Offset Correction: v_off_ini = 232.453 uV, v_off_fin = 234.768 uV 18:12:06 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 93.532 uVs 18:12:06 Integrated Voltage Measurement... 18:12:15 Voltage Offset Correction: v_off_ini = 213.464 uV, v_off_fin = 209.459 uV 18:12:15 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -85.381 uVs 18:12:15 Integrated Voltage Measurement... 18:12:24 Voltage Offset Correction: v_off_ini = 222.015 uV, v_off_fin = 223.160 uV 18:12:24 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 82.152 uVs 18:12:27 I2X Average and RMS of Measurements VT = -91.792 +- 5.686 uVs I2X = 8.445 +- 0.523 uTm^2 18:12:29 Wire Position: X0 = 0.0000 mm, Y0 = -0.5000 mm 18:12:32 Integrated Voltage Measurement... 18:12:41 Voltage Offset Correction: v_off_ini = 232.222 uV, v_off_fin = 223.806 uV 18:12:41 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -297.036 uVs 18:12:41 Integrated Voltage Measurement... 18:12:50 Voltage Offset Correction: v_off_ini = 225.459 uV, v_off_fin = 226.598 uV 18:12:50 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 306.607 uVs 18:12:50 Integrated Voltage Measurement... 18:12:59 Voltage Offset Correction: v_off_ini = 229.066 uV, v_off_fin = 215.662 uV 18:12:59 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -293.214 uVs 18:12:59 Integrated Voltage Measurement... 18:13:08 Voltage Offset Correction: v_off_ini = 228.092 uV, v_off_fin = 221.218 uV 18:13:08 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 300.810 uVs 18:13:08 Integrated Voltage Measurement... 18:13:17 Voltage Offset Correction: v_off_ini = 229.667 uV, v_off_fin = 223.361 uV 18:13:17 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -307.091 uVs 18:13:17 Integrated Voltage Measurement... 18:13:26 Voltage Offset Correction: v_off_ini = 222.672 uV, v_off_fin = 230.290 uV 18:13:26 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 299.354 uVs 18:13:28 I2Y Average and RMS of Measurements VT = -300.685 +- 2.660 uVs I2Y = -27.663 +- 0.245 uTm^2 18:13:32 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 18:13:36 Integrated Voltage Measurement... 18:13:45 Voltage Offset Correction: v_off_ini = 244.519 uV, v_off_fin = 235.392 uV 18:13:45 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -79.073 uVs 18:13:45 Integrated Voltage Measurement... 18:13:54 Voltage Offset Correction: v_off_ini = 229.535 uV, v_off_fin = 243.023 uV 18:13:54 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 82.803 uVs 18:13:54 Integrated Voltage Measurement... 18:14:03 Voltage Offset Correction: v_off_ini = 232.091 uV, v_off_fin = 230.703 uV 18:14:03 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -83.366 uVs 18:14:03 Integrated Voltage Measurement... 18:14:12 Voltage Offset Correction: v_off_ini = 245.598 uV, v_off_fin = 256.796 uV 18:14:12 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.4999 mm dXY = 0.4999 mm VT = 66.989 uVs 18:14:12 Integrated Voltage Measurement... 18:14:21 Voltage Offset Correction: v_off_ini = 244.523 uV, v_off_fin = 237.006 uV 18:14:21 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.4999 mm dXY = -0.4999 mm VT = -82.523 uVs 18:14:21 Integrated Voltage Measurement... 18:14:30 Voltage Offset Correction: v_off_ini = 235.246 uV, v_off_fin = 244.582 uV 18:14:30 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = 88.488 uVs 18:14:35 I1X Average and RMS of Measurements VT = -80.540 +- 4.226 uVs I1X = 16.110 +- 0.845 uTm 18:14:37 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 18:14:41 Integrated Voltage Measurement... 18:14:50 Voltage Offset Correction: v_off_ini = 226.432 uV, v_off_fin = 228.257 uV 18:14:50 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -180.470 uVs 18:14:50 Integrated Voltage Measurement... 18:14:59 Voltage Offset Correction: v_off_ini = 223.266 uV, v_off_fin = 238.264 uV 18:14:59 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 188.336 uVs 18:14:59 Integrated Voltage Measurement... 18:15:08 Voltage Offset Correction: v_off_ini = 211.384 uV, v_off_fin = 198.254 uV 18:15:08 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -195.518 uVs 18:15:08 Integrated Voltage Measurement... 18:15:17 Voltage Offset Correction: v_off_ini = 212.909 uV, v_off_fin = 213.603 uV 18:15:17 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 191.067 uVs 18:15:17 Integrated Voltage Measurement... 18:15:26 Voltage Offset Correction: v_off_ini = 217.382 uV, v_off_fin = 207.893 uV 18:15:26 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -197.027 uVs 18:15:26 Integrated Voltage Measurement... 18:15:35 Voltage Offset Correction: v_off_ini = 212.810 uV, v_off_fin = 224.799 uV 18:15:35 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 189.970 uVs 18:15:39 I1Y Average and RMS of Measurements VT = -190.398 +- 4.240 uVs I1Y = -38.080 +- 0.848 uTm 18:15:42 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 18:15:44 Integrated Voltage Measurement... 18:15:53 Voltage Offset Correction: v_off_ini = 218.550 uV, v_off_fin = 213.103 uV 18:15:53 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -97.078 uVs 18:15:53 Integrated Voltage Measurement... 18:16:02 Voltage Offset Correction: v_off_ini = 207.141 uV, v_off_fin = 211.085 uV 18:16:02 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 103.079 uVs 18:16:02 Integrated Voltage Measurement... 18:16:11 Voltage Offset Correction: v_off_ini = 213.397 uV, v_off_fin = 204.528 uV 18:16:11 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -92.363 uVs 18:16:11 Integrated Voltage Measurement... 18:16:20 Voltage Offset Correction: v_off_ini = 200.278 uV, v_off_fin = 221.512 uV 18:16:20 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 89.183 uVs 18:16:20 Integrated Voltage Measurement... 18:16:29 Voltage Offset Correction: v_off_ini = 216.906 uV, v_off_fin = 209.538 uV 18:16:29 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -80.822 uVs 18:16:29 Integrated Voltage Measurement... 18:16:38 Voltage Offset Correction: v_off_ini = 205.227 uV, v_off_fin = 217.378 uV 18:16:38 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 98.478 uVs 18:16:41 I2X Average and RMS of Measurements VT = -93.501 +- 4.674 uVs I2X = 8.602 +- 0.430 uTm^2 18:16:43 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 18:16:46 Integrated Voltage Measurement... 18:16:55 Voltage Offset Correction: v_off_ini = 225.940 uV, v_off_fin = 220.497 uV 18:16:55 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -296.540 uVs 18:16:55 Integrated Voltage Measurement... 18:17:04 Voltage Offset Correction: v_off_ini = 200.844 uV, v_off_fin = 194.169 uV 18:17:04 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 309.945 uVs 18:17:04 Integrated Voltage Measurement... 18:17:13 Voltage Offset Correction: v_off_ini = 195.060 uV, v_off_fin = 197.757 uV 18:17:13 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -312.910 uVs 18:17:13 Integrated Voltage Measurement... 18:17:22 Voltage Offset Correction: v_off_ini = 195.626 uV, v_off_fin = 201.401 uV 18:17:22 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 294.767 uVs 18:17:22 Integrated Voltage Measurement... 18:17:31 Voltage Offset Correction: v_off_ini = 205.625 uV, v_off_fin = 208.005 uV 18:17:31 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -305.624 uVs 18:17:31 Integrated Voltage Measurement... 18:17:40 Voltage Offset Correction: v_off_ini = 213.018 uV, v_off_fin = 215.658 uV 18:17:40 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 305.114 uVs 18:17:42 I2Y Average and RMS of Measurements VT = -304.150 +- 0.896 uVs I2Y = -27.982 +- 0.082 uTm^2 18:17:46 Wire Position: X0 = 0.0000 mm, Y0 = 0.5000 mm 18:17:51 Integrated Voltage Measurement... 18:18:00 Voltage Offset Correction: v_off_ini = 217.592 uV, v_off_fin = 218.223 uV 18:18:00 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -81.954 uVs 18:18:00 Integrated Voltage Measurement... 18:18:09 Voltage Offset Correction: v_off_ini = 218.857 uV, v_off_fin = 225.337 uV 18:18:09 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 68.332 uVs 18:18:09 Integrated Voltage Measurement... 18:18:18 Voltage Offset Correction: v_off_ini = 230.120 uV, v_off_fin = 218.624 uV 18:18:18 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -86.872 uVs 18:18:18 Integrated Voltage Measurement... 18:18:27 Voltage Offset Correction: v_off_ini = 221.069 uV, v_off_fin = 225.327 uV 18:18:27 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 49.953 uVs 18:18:27 Integrated Voltage Measurement... 18:18:36 Voltage Offset Correction: v_off_ini = 215.856 uV, v_off_fin = 207.950 uV 18:18:36 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -61.491 uVs 18:18:36 Integrated Voltage Measurement... 18:18:45 Voltage Offset Correction: v_off_ini = 225.013 uV, v_off_fin = 218.920 uV 18:18:45 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 88.831 uVs 18:18:49 I1X Average and RMS of Measurements VT = -72.906 +- 3.177 uVs I1X = 14.582 +- 0.635 uTm 18:18:51 Wire Position: X0 = 0.0000 mm, Y0 = 0.5000 mm 18:18:56 Integrated Voltage Measurement... 18:19:04 Voltage Offset Correction: v_off_ini = 230.850 uV, v_off_fin = 218.773 uV 18:19:04 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -193.670 uVs 18:19:05 Integrated Voltage Measurement... 18:19:13 Voltage Offset Correction: v_off_ini = 233.473 uV, v_off_fin = 252.545 uV 18:19:13 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 195.126 uVs 18:19:13 Integrated Voltage Measurement... 18:19:22 Voltage Offset Correction: v_off_ini = 237.985 uV, v_off_fin = 221.126 uV 18:19:22 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -171.515 uVs 18:19:22 Integrated Voltage Measurement... 18:19:31 Voltage Offset Correction: v_off_ini = 228.062 uV, v_off_fin = 228.564 uV 18:19:31 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 195.060 uVs 18:19:31 Integrated Voltage Measurement... 18:19:40 Voltage Offset Correction: v_off_ini = 228.121 uV, v_off_fin = 209.613 uV 18:19:40 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -187.482 uVs 18:19:40 Integrated Voltage Measurement... 18:19:49 Voltage Offset Correction: v_off_ini = 216.299 uV, v_off_fin = 228.623 uV 18:19:49 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 197.142 uVs 18:19:54 I1Y Average and RMS of Measurements VT = -189.999 +- 4.822 uVs I1Y = -37.997 +- 0.964 uTm 18:19:56 Wire Position: X0 = 0.0000 mm, Y0 = 0.5000 mm 18:19:59 Integrated Voltage Measurement... 18:20:08 Voltage Offset Correction: v_off_ini = 220.130 uV, v_off_fin = 220.145 uV 18:20:08 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -81.321 uVs 18:20:08 Integrated Voltage Measurement... 18:20:17 Voltage Offset Correction: v_off_ini = 226.003 uV, v_off_fin = 233.112 uV 18:20:17 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 76.048 uVs 18:20:17 Integrated Voltage Measurement... 18:20:26 Voltage Offset Correction: v_off_ini = 225.004 uV, v_off_fin = 216.252 uV 18:20:26 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -101.039 uVs 18:20:26 Integrated Voltage Measurement... 18:20:35 Voltage Offset Correction: v_off_ini = 230.469 uV, v_off_fin = 248.782 uV 18:20:35 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 70.972 uVs 18:20:35 Integrated Voltage Measurement... 18:20:44 Voltage Offset Correction: v_off_ini = 228.190 uV, v_off_fin = 215.447 uV 18:20:44 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -88.034 uVs 18:20:44 Integrated Voltage Measurement... 18:20:53 Voltage Offset Correction: v_off_ini = 228.900 uV, v_off_fin = 246.137 uV 18:20:53 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 100.246 uVs 18:20:55 I2X Average and RMS of Measurements VT = -86.277 +- 6.313 uVs I2X = 7.938 +- 0.581 uTm^2 18:20:57 Wire Position: X0 = 0.0000 mm, Y0 = 0.5000 mm 18:21:00 Integrated Voltage Measurement... 18:21:09 Voltage Offset Correction: v_off_ini = 234.479 uV, v_off_fin = 228.476 uV 18:21:09 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -291.987 uVs 18:21:09 Integrated Voltage Measurement... 18:21:18 Voltage Offset Correction: v_off_ini = 229.802 uV, v_off_fin = 237.440 uV 18:21:18 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 312.446 uVs 18:21:18 Integrated Voltage Measurement... 18:21:27 Voltage Offset Correction: v_off_ini = 233.960 uV, v_off_fin = 234.304 uV 18:21:27 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -307.404 uVs 18:21:27 Integrated Voltage Measurement... 18:21:36 Voltage Offset Correction: v_off_ini = 232.006 uV, v_off_fin = 243.055 uV 18:21:36 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 290.958 uVs 18:21:36 Integrated Voltage Measurement... 18:21:45 Voltage Offset Correction: v_off_ini = 227.583 uV, v_off_fin = 228.203 uV 18:21:45 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -305.617 uVs 18:21:45 Integrated Voltage Measurement... 18:21:54 Voltage Offset Correction: v_off_ini = 235.976 uV, v_off_fin = 229.768 uV 18:21:54 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 290.746 uVs 18:21:57 I2Y Average and RMS of Measurements VT = -299.860 +- 1.716 uVs I2Y = -27.587 +- 0.158 uTm^2 18:22:01 Wire Position: X0 = 0.0000 mm, Y0 = 1.0000 mm 18:22:05 Integrated Voltage Measurement... 18:22:14 Voltage Offset Correction: v_off_ini = 247.330 uV, v_off_fin = 246.388 uV 18:22:14 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -72.111 uVs 18:22:14 Integrated Voltage Measurement... 18:22:23 Voltage Offset Correction: v_off_ini = 243.108 uV, v_off_fin = 238.340 uV 18:22:23 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 68.851 uVs 18:22:23 Integrated Voltage Measurement... 18:22:32 Voltage Offset Correction: v_off_ini = 243.455 uV, v_off_fin = 235.427 uV 18:22:32 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -67.165 uVs 18:22:32 Integrated Voltage Measurement... 18:22:41 Voltage Offset Correction: v_off_ini = 239.439 uV, v_off_fin = 234.582 uV 18:22:41 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 76.420 uVs 18:22:41 Integrated Voltage Measurement... 18:22:50 Voltage Offset Correction: v_off_ini = 245.521 uV, v_off_fin = 242.136 uV 18:22:50 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -92.675 uVs 18:22:50 Integrated Voltage Measurement... 18:22:59 Voltage Offset Correction: v_off_ini = 227.003 uV, v_off_fin = 241.791 uV 18:22:59 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 76.031 uVs 18:23:03 I1X Average and RMS of Measurements VT = -75.542 +- 6.253 uVs I1X = 15.108 +- 1.251 uTm 18:23:06 Wire Position: X0 = 0.0000 mm, Y0 = 1.0000 mm 18:23:10 Integrated Voltage Measurement... 18:23:19 Voltage Offset Correction: v_off_ini = 226.728 uV, v_off_fin = 214.195 uV 18:23:19 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -162.501 uVs 18:23:19 Integrated Voltage Measurement... 18:23:28 Voltage Offset Correction: v_off_ini = 229.008 uV, v_off_fin = 244.203 uV 18:23:28 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 184.084 uVs 18:23:28 Integrated Voltage Measurement... 18:23:37 Voltage Offset Correction: v_off_ini = 221.817 uV, v_off_fin = 221.169 uV 18:23:37 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -184.430 uVs 18:23:37 Integrated Voltage Measurement... 18:23:46 Voltage Offset Correction: v_off_ini = 237.386 uV, v_off_fin = 238.422 uV 18:23:46 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 175.705 uVs 18:23:46 Integrated Voltage Measurement... 18:23:55 Voltage Offset Correction: v_off_ini = 236.715 uV, v_off_fin = 229.325 uV 18:23:55 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -182.711 uVs 18:23:55 Integrated Voltage Measurement... 18:24:04 Voltage Offset Correction: v_off_ini = 227.518 uV, v_off_fin = 244.982 uV 18:24:04 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 187.508 uVs 18:24:08 I1Y Average and RMS of Measurements VT = -179.490 +- 4.842 uVs I1Y = -35.896 +- 0.967 uTm 18:24:10 Wire Position: X0 = 0.0000 mm, Y0 = 1.0000 mm 18:24:13 Integrated Voltage Measurement... 18:24:22 Voltage Offset Correction: v_off_ini = 229.082 uV, v_off_fin = 210.464 uV 18:24:22 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -71.868 uVs 18:24:22 Integrated Voltage Measurement... 18:24:31 Voltage Offset Correction: v_off_ini = 228.509 uV, v_off_fin = 239.489 uV 18:24:31 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 77.482 uVs 18:24:31 Integrated Voltage Measurement... 18:24:40 Voltage Offset Correction: v_off_ini = 225.542 uV, v_off_fin = 213.877 uV 18:24:40 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -88.623 uVs 18:24:40 Integrated Voltage Measurement... 18:24:49 Voltage Offset Correction: v_off_ini = 222.899 uV, v_off_fin = 234.045 uV 18:24:49 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 70.813 uVs 18:24:49 Integrated Voltage Measurement... 18:24:58 Voltage Offset Correction: v_off_ini = 218.556 uV, v_off_fin = 231.514 uV 18:24:58 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -86.796 uVs 18:24:58 Integrated Voltage Measurement... 18:25:07 Voltage Offset Correction: v_off_ini = 233.327 uV, v_off_fin = 228.802 uV 18:25:07 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 75.653 uVs 18:25:09 I2X Average and RMS of Measurements VT = -78.539 +- 2.801 uVs I2X = 7.226 +- 0.258 uTm^2 18:25:12 Wire Position: X0 = 0.0000 mm, Y0 = 1.0000 mm 18:25:14 Integrated Voltage Measurement... 18:25:23 Voltage Offset Correction: v_off_ini = 232.088 uV, v_off_fin = 218.986 uV 18:25:23 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -302.544 uVs 18:25:23 Integrated Voltage Measurement... 18:25:32 Voltage Offset Correction: v_off_ini = 215.465 uV, v_off_fin = 226.349 uV 18:25:32 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 297.180 uVs 18:25:32 Integrated Voltage Measurement... 18:25:41 Voltage Offset Correction: v_off_ini = 222.887 uV, v_off_fin = 226.931 uV 18:25:41 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -284.883 uVs 18:25:41 Integrated Voltage Measurement... 18:25:50 Voltage Offset Correction: v_off_ini = 236.422 uV, v_off_fin = 227.676 uV 18:25:50 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 290.083 uVs 18:25:50 Integrated Voltage Measurement... 18:25:59 Voltage Offset Correction: v_off_ini = 231.640 uV, v_off_fin = 239.912 uV 18:25:59 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -296.146 uVs 18:25:59 Integrated Voltage Measurement... 18:26:08 Voltage Offset Correction: v_off_ini = 247.119 uV, v_off_fin = 252.065 uV 18:26:08 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 284.446 uVs 18:26:11 I2Y Average and RMS of Measurements VT = -292.547 +- 5.298 uVs I2Y = -26.914 +- 0.487 uTm^2 18:26:28 Wire Position: X0 = -2.0000 mm, Y0 = 0.0000 mm 18:26:32 Integrated Voltage Measurement... 18:26:41 Voltage Offset Correction: v_off_ini = 240.963 uV, v_off_fin = 233.843 uV 18:26:41 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.4999 mm dXY = -0.4999 mm VT = -117.178 uVs 18:26:41 Integrated Voltage Measurement... 18:26:50 Voltage Offset Correction: v_off_ini = 233.870 uV, v_off_fin = 257.886 uV 18:26:50 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 95.812 uVs 18:26:50 Integrated Voltage Measurement... 18:26:59 Voltage Offset Correction: v_off_ini = 241.335 uV, v_off_fin = 250.397 uV 18:26:59 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -115.523 uVs 18:26:59 Integrated Voltage Measurement... 18:27:08 Voltage Offset Correction: v_off_ini = 244.380 uV, v_off_fin = 257.413 uV 18:27:08 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 100.323 uVs 18:27:08 Integrated Voltage Measurement... 18:27:17 Voltage Offset Correction: v_off_ini = 245.584 uV, v_off_fin = 235.889 uV 18:27:17 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -118.880 uVs 18:27:17 Integrated Voltage Measurement... 18:27:26 Voltage Offset Correction: v_off_ini = 223.923 uV, v_off_fin = 242.161 uV 18:27:26 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 123.879 uVs 18:27:30 I1X Average and RMS of Measurements VT = -111.933 +- 6.705 uVs I1X = 22.387 +- 1.340 uTm 18:27:33 Wire Position: X0 = -2.0000 mm, Y0 = 0.0000 mm 18:27:37 Integrated Voltage Measurement... 18:27:46 Voltage Offset Correction: v_off_ini = 249.566 uV, v_off_fin = 229.943 uV 18:27:46 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -203.230 uVs 18:27:46 Integrated Voltage Measurement... 18:27:55 Voltage Offset Correction: v_off_ini = 246.945 uV, v_off_fin = 251.884 uV 18:27:55 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 179.985 uVs 18:27:55 Integrated Voltage Measurement... 18:28:04 Voltage Offset Correction: v_off_ini = 248.864 uV, v_off_fin = 232.978 uV 18:28:04 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -183.641 uVs 18:28:04 Integrated Voltage Measurement... 18:28:13 Voltage Offset Correction: v_off_ini = 244.444 uV, v_off_fin = 265.303 uV 18:28:13 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 175.103 uVs 18:28:13 Integrated Voltage Measurement... 18:28:22 Voltage Offset Correction: v_off_ini = 237.941 uV, v_off_fin = 228.904 uV 18:28:22 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -186.900 uVs 18:28:22 Integrated Voltage Measurement... 18:28:31 Voltage Offset Correction: v_off_ini = 246.873 uV, v_off_fin = 258.675 uV 18:28:31 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 171.082 uVs 18:28:35 I1Y Average and RMS of Measurements VT = -183.324 +- 5.860 uVs I1Y = -36.665 +- 1.172 uTm 18:28:40 Wire Position: X0 = -2.0000 mm, Y0 = 0.0000 mm 18:28:43 Integrated Voltage Measurement... 18:28:52 Voltage Offset Correction: v_off_ini = 240.897 uV, v_off_fin = 208.763 uV 18:28:52 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -136.136 uVs 18:28:52 Integrated Voltage Measurement... 18:29:01 Voltage Offset Correction: v_off_ini = 233.313 uV, v_off_fin = 241.903 uV 18:29:01 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 115.397 uVs 18:29:01 Integrated Voltage Measurement... 18:29:10 Voltage Offset Correction: v_off_ini = 232.866 uV, v_off_fin = 218.049 uV 18:29:10 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -115.243 uVs 18:29:10 Integrated Voltage Measurement... 18:29:19 Voltage Offset Correction: v_off_ini = 219.561 uV, v_off_fin = 233.602 uV 18:29:19 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 111.929 uVs 18:29:19 Integrated Voltage Measurement... 18:29:28 Voltage Offset Correction: v_off_ini = 225.292 uV, v_off_fin = 218.017 uV 18:29:28 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -108.504 uVs 18:29:28 Integrated Voltage Measurement... 18:29:37 Voltage Offset Correction: v_off_ini = 229.519 uV, v_off_fin = 231.532 uV 18:29:37 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 127.251 uVs 18:29:39 I2X Average and RMS of Measurements VT = -119.076 +- 5.044 uVs I2X = 10.955 +- 0.464 uTm^2 18:29:44 Wire Position: X0 = -2.0000 mm, Y0 = 0.0000 mm 18:29:47 Integrated Voltage Measurement... 18:29:56 Voltage Offset Correction: v_off_ini = 210.250 uV, v_off_fin = 214.022 uV 18:29:56 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -282.620 uVs 18:29:56 Integrated Voltage Measurement... 18:30:05 Voltage Offset Correction: v_off_ini = 230.324 uV, v_off_fin = 225.806 uV 18:30:05 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 279.760 uVs 18:30:05 Integrated Voltage Measurement... 18:30:14 Voltage Offset Correction: v_off_ini = 231.373 uV, v_off_fin = 229.144 uV 18:30:14 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -280.341 uVs 18:30:14 Integrated Voltage Measurement... 18:30:23 Voltage Offset Correction: v_off_ini = 233.116 uV, v_off_fin = 229.544 uV 18:30:23 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 269.770 uVs 18:30:23 Integrated Voltage Measurement... 18:30:32 Voltage Offset Correction: v_off_ini = 238.871 uV, v_off_fin = 230.366 uV 18:30:32 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -285.899 uVs 18:30:32 Integrated Voltage Measurement... 18:30:41 Voltage Offset Correction: v_off_ini = 244.932 uV, v_off_fin = 235.832 uV 18:30:41 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 269.620 uVs 18:30:43 I2Y Average and RMS of Measurements VT = -278.002 +- 2.510 uVs I2Y = -25.576 +- 0.231 uTm^2 18:30:48 Wire Position: X0 = -1.0000 mm, Y0 = 0.0000 mm 18:30:52 Integrated Voltage Measurement... 18:31:01 Voltage Offset Correction: v_off_ini = 224.621 uV, v_off_fin = 237.031 uV 18:31:01 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -116.516 uVs 18:31:01 Integrated Voltage Measurement... 18:31:10 Voltage Offset Correction: v_off_ini = 244.843 uV, v_off_fin = 249.680 uV 18:31:10 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 101.213 uVs 18:31:10 Integrated Voltage Measurement... 18:31:19 Voltage Offset Correction: v_off_ini = 228.186 uV, v_off_fin = 214.579 uV 18:31:19 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -98.576 uVs 18:31:19 Integrated Voltage Measurement... 18:31:28 Voltage Offset Correction: v_off_ini = 226.853 uV, v_off_fin = 236.509 uV 18:31:28 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 102.332 uVs 18:31:28 Integrated Voltage Measurement... 18:31:37 Voltage Offset Correction: v_off_ini = 226.052 uV, v_off_fin = 224.455 uV 18:31:37 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -96.663 uVs 18:31:37 Integrated Voltage Measurement... 18:31:46 Voltage Offset Correction: v_off_ini = 221.450 uV, v_off_fin = 222.038 uV 18:31:46 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 108.034 uVs 18:31:50 I1X Average and RMS of Measurements VT = -103.889 +- 3.602 uVs I1X = 20.778 +- 0.721 uTm 18:31:53 Wire Position: X0 = -1.0000 mm, Y0 = 0.0000 mm 18:31:57 Integrated Voltage Measurement... 18:32:06 Voltage Offset Correction: v_off_ini = 221.652 uV, v_off_fin = 214.703 uV 18:32:06 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -189.294 uVs 18:32:06 Integrated Voltage Measurement... 18:32:15 Voltage Offset Correction: v_off_ini = 213.315 uV, v_off_fin = 232.757 uV 18:32:15 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 198.236 uVs 18:32:15 Integrated Voltage Measurement... 18:32:24 Voltage Offset Correction: v_off_ini = 220.034 uV, v_off_fin = 212.513 uV 18:32:24 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -193.315 uVs 18:32:24 Integrated Voltage Measurement... 18:32:33 Voltage Offset Correction: v_off_ini = 216.579 uV, v_off_fin = 239.081 uV 18:32:33 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 191.377 uVs 18:32:33 Integrated Voltage Measurement... 18:32:42 Voltage Offset Correction: v_off_ini = 223.560 uV, v_off_fin = 216.263 uV 18:32:42 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -193.579 uVs 18:32:42 Integrated Voltage Measurement... 18:32:51 Voltage Offset Correction: v_off_ini = 242.725 uV, v_off_fin = 253.088 uV 18:32:51 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 186.615 uVs 18:32:55 I1Y Average and RMS of Measurements VT = -192.069 +- 1.510 uVs I1Y = -38.414 +- 0.302 uTm 18:33:00 Wire Position: X0 = -1.0000 mm, Y0 = 0.0000 mm 18:33:03 Integrated Voltage Measurement... 18:33:12 Voltage Offset Correction: v_off_ini = 211.888 uV, v_off_fin = 207.716 uV 18:33:12 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -102.322 uVs 18:33:12 Integrated Voltage Measurement... 18:33:21 Voltage Offset Correction: v_off_ini = 227.252 uV, v_off_fin = 239.022 uV 18:33:21 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 106.039 uVs 18:33:21 Integrated Voltage Measurement... 18:33:30 Voltage Offset Correction: v_off_ini = 217.982 uV, v_off_fin = 216.538 uV 18:33:30 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -99.260 uVs 18:33:30 Integrated Voltage Measurement... 18:33:39 Voltage Offset Correction: v_off_ini = 228.622 uV, v_off_fin = 240.383 uV 18:33:39 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 122.910 uVs 18:33:39 Integrated Voltage Measurement... 18:33:48 Voltage Offset Correction: v_off_ini = 235.185 uV, v_off_fin = 225.806 uV 18:33:48 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -117.551 uVs 18:33:48 Integrated Voltage Measurement... 18:33:57 Voltage Offset Correction: v_off_ini = 227.181 uV, v_off_fin = 242.653 uV 18:33:57 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 109.354 uVs 18:33:59 I2X Average and RMS of Measurements VT = -109.572 +- 3.933 uVs I2X = 10.081 +- 0.362 uTm^2 18:34:04 Wire Position: X0 = -1.0000 mm, Y0 = 0.0000 mm 18:34:07 Integrated Voltage Measurement... 18:34:16 Voltage Offset Correction: v_off_ini = 240.317 uV, v_off_fin = 234.250 uV 18:34:16 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -275.651 uVs 18:34:16 Integrated Voltage Measurement... 18:34:25 Voltage Offset Correction: v_off_ini = 223.403 uV, v_off_fin = 230.064 uV 18:34:25 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 286.850 uVs 18:34:25 Integrated Voltage Measurement... 18:34:34 Voltage Offset Correction: v_off_ini = 235.518 uV, v_off_fin = 230.313 uV 18:34:34 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -302.028 uVs 18:34:34 Integrated Voltage Measurement... 18:34:43 Voltage Offset Correction: v_off_ini = 242.089 uV, v_off_fin = 226.420 uV 18:34:43 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 287.444 uVs 18:34:43 Integrated Voltage Measurement... 18:34:52 Voltage Offset Correction: v_off_ini = 219.909 uV, v_off_fin = 219.158 uV 18:34:52 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -288.856 uVs 18:34:52 Integrated Voltage Measurement... 18:35:01 Voltage Offset Correction: v_off_ini = 223.287 uV, v_off_fin = 226.698 uV 18:35:01 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 278.176 uVs 18:35:03 I2Y Average and RMS of Measurements VT = -286.501 +- 5.896 uVs I2Y = -26.358 +- 0.542 uTm^2 18:35:08 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 18:35:12 Integrated Voltage Measurement... 18:35:21 Voltage Offset Correction: v_off_ini = 234.592 uV, v_off_fin = 233.410 uV 18:35:21 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -111.795 uVs 18:35:21 Integrated Voltage Measurement... 18:35:30 Voltage Offset Correction: v_off_ini = 226.112 uV, v_off_fin = 243.657 uV 18:35:30 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 92.409 uVs 18:35:30 Integrated Voltage Measurement... 18:35:39 Voltage Offset Correction: v_off_ini = 231.263 uV, v_off_fin = 231.146 uV 18:35:39 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -90.724 uVs 18:35:39 Integrated Voltage Measurement... 18:35:48 Voltage Offset Correction: v_off_ini = 242.988 uV, v_off_fin = 240.574 uV 18:35:48 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 81.180 uVs 18:35:48 Integrated Voltage Measurement... 18:35:57 Voltage Offset Correction: v_off_ini = 223.074 uV, v_off_fin = 228.925 uV 18:35:57 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -82.883 uVs 18:35:57 Integrated Voltage Measurement... 18:36:06 Voltage Offset Correction: v_off_ini = 227.349 uV, v_off_fin = 219.944 uV 18:36:06 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 77.559 uVs 18:36:10 I1X Average and RMS of Measurements VT = -89.425 +- 9.264 uVs I1X = 17.886 +- 1.852 uTm 18:36:13 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 18:36:17 Integrated Voltage Measurement... 18:36:26 Voltage Offset Correction: v_off_ini = 213.295 uV, v_off_fin = 210.860 uV 18:36:26 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -188.132 uVs 18:36:26 Integrated Voltage Measurement... 18:36:35 Voltage Offset Correction: v_off_ini = 221.729 uV, v_off_fin = 239.292 uV 18:36:35 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 198.828 uVs 18:36:35 Integrated Voltage Measurement... 18:36:44 Voltage Offset Correction: v_off_ini = 213.668 uV, v_off_fin = 214.353 uV 18:36:44 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -189.676 uVs 18:36:44 Integrated Voltage Measurement... 18:36:53 Voltage Offset Correction: v_off_ini = 218.492 uV, v_off_fin = 235.323 uV 18:36:53 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 189.747 uVs 18:36:53 Integrated Voltage Measurement... 18:37:02 Voltage Offset Correction: v_off_ini = 229.427 uV, v_off_fin = 224.533 uV 18:37:02 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -191.350 uVs 18:37:02 Integrated Voltage Measurement... 18:37:11 Voltage Offset Correction: v_off_ini = 213.985 uV, v_off_fin = 230.017 uV 18:37:11 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 178.222 uVs 18:37:15 I1Y Average and RMS of Measurements VT = -189.326 +- 3.560 uVs I1Y = -37.865 +- 0.712 uTm 18:37:20 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 18:37:23 Integrated Voltage Measurement... 18:37:32 Voltage Offset Correction: v_off_ini = 202.262 uV, v_off_fin = 203.544 uV 18:37:32 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -107.002 uVs 18:37:32 Integrated Voltage Measurement... 18:37:41 Voltage Offset Correction: v_off_ini = 216.375 uV, v_off_fin = 220.237 uV 18:37:41 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 122.169 uVs 18:37:41 Integrated Voltage Measurement... 18:37:50 Voltage Offset Correction: v_off_ini = 228.177 uV, v_off_fin = 222.925 uV 18:37:50 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -103.314 uVs 18:37:50 Integrated Voltage Measurement... 18:37:59 Voltage Offset Correction: v_off_ini = 232.133 uV, v_off_fin = 232.974 uV 18:37:59 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 111.154 uVs 18:37:59 Integrated Voltage Measurement... 18:38:08 Voltage Offset Correction: v_off_ini = 207.765 uV, v_off_fin = 213.545 uV 18:38:08 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -97.375 uVs 18:38:08 Integrated Voltage Measurement... 18:38:17 Voltage Offset Correction: v_off_ini = 217.860 uV, v_off_fin = 223.445 uV 18:38:17 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 126.191 uVs 18:38:20 I2X Average and RMS of Measurements VT = -111.201 +- 3.029 uVs I2X = 10.230 +- 0.279 uTm^2 18:38:25 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 18:38:27 Integrated Voltage Measurement... 18:38:36 Voltage Offset Correction: v_off_ini = 231.330 uV, v_off_fin = 230.493 uV 18:38:36 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -273.953 uVs 18:38:36 Integrated Voltage Measurement... 18:38:45 Voltage Offset Correction: v_off_ini = 231.280 uV, v_off_fin = 233.134 uV 18:38:45 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 295.371 uVs 18:38:45 Integrated Voltage Measurement... 18:38:54 Voltage Offset Correction: v_off_ini = 230.708 uV, v_off_fin = 225.707 uV 18:38:54 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -290.505 uVs 18:38:54 Integrated Voltage Measurement... 18:39:03 Voltage Offset Correction: v_off_ini = 245.343 uV, v_off_fin = 234.089 uV 18:39:03 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 290.355 uVs 18:39:03 Integrated Voltage Measurement... 18:39:12 Voltage Offset Correction: v_off_ini = 239.614 uV, v_off_fin = 227.604 uV 18:39:12 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -283.915 uVs 18:39:12 Integrated Voltage Measurement... 18:39:21 Voltage Offset Correction: v_off_ini = 234.916 uV, v_off_fin = 232.896 uV 18:39:21 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 273.845 uVs 18:39:24 I2Y Average and RMS of Measurements VT = -284.658 +- 4.715 uVs I2Y = -26.188 +- 0.434 uTm^2 18:39:28 Wire Position: X0 = 1.0000 mm, Y0 = 0.0000 mm 18:39:33 Integrated Voltage Measurement... 18:39:42 Voltage Offset Correction: v_off_ini = 229.584 uV, v_off_fin = 222.696 uV 18:39:42 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -89.945 uVs 18:39:42 Integrated Voltage Measurement... 18:39:51 Voltage Offset Correction: v_off_ini = 230.404 uV, v_off_fin = 238.910 uV 18:39:51 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 93.348 uVs 18:39:51 Integrated Voltage Measurement... 18:40:00 Voltage Offset Correction: v_off_ini = 227.024 uV, v_off_fin = 209.814 uV 18:40:00 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -99.642 uVs 18:40:00 Integrated Voltage Measurement... 18:40:09 Voltage Offset Correction: v_off_ini = 217.387 uV, v_off_fin = 217.735 uV 18:40:09 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 100.478 uVs 18:40:09 Integrated Voltage Measurement... 18:40:18 Voltage Offset Correction: v_off_ini = 230.899 uV, v_off_fin = 235.727 uV 18:40:18 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -92.710 uVs 18:40:18 Integrated Voltage Measurement... 18:40:27 Voltage Offset Correction: v_off_ini = 221.565 uV, v_off_fin = 224.826 uV 18:40:27 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 88.376 uVs 18:40:31 I1X Average and RMS of Measurements VT = -94.083 +- 4.250 uVs I1X = 18.817 +- 0.850 uTm 18:40:33 Wire Position: X0 = 1.0000 mm, Y0 = 0.0000 mm 18:40:38 Integrated Voltage Measurement... 18:40:47 Voltage Offset Correction: v_off_ini = 208.650 uV, v_off_fin = 203.984 uV 18:40:47 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -198.047 uVs 18:40:47 Integrated Voltage Measurement... 18:40:56 Voltage Offset Correction: v_off_ini = 224.783 uV, v_off_fin = 220.587 uV 18:40:56 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 181.613 uVs 18:40:56 Integrated Voltage Measurement... 18:41:05 Voltage Offset Correction: v_off_ini = 213.662 uV, v_off_fin = 215.286 uV 18:41:05 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -175.592 uVs 18:41:05 Integrated Voltage Measurement... 18:41:13 Voltage Offset Correction: v_off_ini = 215.679 uV, v_off_fin = 228.043 uV 18:41:14 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 182.193 uVs 18:41:14 Integrated Voltage Measurement... 18:41:22 Voltage Offset Correction: v_off_ini = 215.685 uV, v_off_fin = 207.271 uV 18:41:23 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -172.557 uVs 18:41:23 Integrated Voltage Measurement... 18:41:31 Voltage Offset Correction: v_off_ini = 202.071 uV, v_off_fin = 219.135 uV 18:41:31 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 181.035 uVs 18:41:36 I1Y Average and RMS of Measurements VT = -181.839 +- 5.714 uVs I1Y = -36.368 +- 1.143 uTm 18:41:41 Wire Position: X0 = 1.0000 mm, Y0 = 0.0000 mm 18:41:43 Integrated Voltage Measurement... 18:41:52 Voltage Offset Correction: v_off_ini = 204.286 uV, v_off_fin = 201.522 uV 18:41:52 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -122.015 uVs 18:41:52 Integrated Voltage Measurement... 18:42:01 Voltage Offset Correction: v_off_ini = 201.905 uV, v_off_fin = 203.537 uV 18:42:01 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 94.504 uVs 18:42:01 Integrated Voltage Measurement... 18:42:10 Voltage Offset Correction: v_off_ini = 213.958 uV, v_off_fin = 201.639 uV 18:42:10 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -110.726 uVs 18:42:11 Integrated Voltage Measurement... 18:42:19 Voltage Offset Correction: v_off_ini = 199.615 uV, v_off_fin = 211.286 uV 18:42:19 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 109.484 uVs 18:42:20 Integrated Voltage Measurement... 18:42:28 Voltage Offset Correction: v_off_ini = 197.070 uV, v_off_fin = 196.012 uV 18:42:29 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -104.579 uVs 18:42:29 Integrated Voltage Measurement... 18:42:38 Voltage Offset Correction: v_off_ini = 208.416 uV, v_off_fin = 216.354 uV 18:42:38 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 118.657 uVs 18:42:40 I2X Average and RMS of Measurements VT = -109.994 +- 1.374 uVs I2X = 10.119 +- 0.126 uTm^2 18:42:45 Wire Position: X0 = 1.0000 mm, Y0 = 0.0000 mm 18:42:48 Integrated Voltage Measurement... 18:42:57 Voltage Offset Correction: v_off_ini = 188.691 uV, v_off_fin = 204.935 uV 18:42:57 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -285.660 uVs 18:42:57 Integrated Voltage Measurement... 18:43:06 Voltage Offset Correction: v_off_ini = 215.001 uV, v_off_fin = 216.119 uV 18:43:06 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 274.477 uVs 18:43:06 Integrated Voltage Measurement... 18:43:15 Voltage Offset Correction: v_off_ini = 203.892 uV, v_off_fin = 197.613 uV 18:43:15 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -294.887 uVs 18:43:15 Integrated Voltage Measurement... 18:43:24 Voltage Offset Correction: v_off_ini = 207.609 uV, v_off_fin = 206.603 uV 18:43:24 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = 277.255 uVs 18:43:24 Integrated Voltage Measurement... 18:43:33 Voltage Offset Correction: v_off_ini = 210.077 uV, v_off_fin = 209.535 uV 18:43:33 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -287.051 uVs 18:43:33 Integrated Voltage Measurement... 18:43:42 Voltage Offset Correction: v_off_ini = 204.741 uV, v_off_fin = 214.136 uV 18:43:42 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 279.601 uVs 18:43:44 I2Y Average and RMS of Measurements VT = -283.155 +- 2.453 uVs I2Y = -26.051 +- 0.226 uTm^2 18:43:49 Wire Position: X0 = 2.0000 mm, Y0 = 0.0000 mm 18:43:53 Integrated Voltage Measurement... 18:44:02 Voltage Offset Correction: v_off_ini = 224.386 uV, v_off_fin = 203.983 uV 18:44:02 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -110.852 uVs 18:44:02 Integrated Voltage Measurement... 18:44:11 Voltage Offset Correction: v_off_ini = 208.750 uV, v_off_fin = 204.068 uV 18:44:11 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 97.719 uVs 18:44:11 Integrated Voltage Measurement... 18:44:20 Voltage Offset Correction: v_off_ini = 207.266 uV, v_off_fin = 206.070 uV 18:44:20 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -86.134 uVs 18:44:20 Integrated Voltage Measurement... 18:44:29 Voltage Offset Correction: v_off_ini = 217.197 uV, v_off_fin = 213.486 uV 18:44:29 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 101.921 uVs 18:44:29 Integrated Voltage Measurement... 18:44:38 Voltage Offset Correction: v_off_ini = 212.296 uV, v_off_fin = 200.065 uV 18:44:38 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -98.859 uVs 18:44:38 Integrated Voltage Measurement... 18:44:47 Voltage Offset Correction: v_off_ini = 210.147 uV, v_off_fin = 204.873 uV 18:44:47 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 102.238 uVs 18:44:51 I1X Average and RMS of Measurements VT = -99.620 +- 4.239 uVs I1X = 19.924 +- 0.848 uTm 18:44:54 Wire Position: X0 = 2.0000 mm, Y0 = 0.0000 mm 18:44:58 Integrated Voltage Measurement... 18:45:07 Voltage Offset Correction: v_off_ini = 210.882 uV, v_off_fin = 193.283 uV 18:45:07 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -185.065 uVs 18:45:07 Integrated Voltage Measurement... 18:45:16 Voltage Offset Correction: v_off_ini = 206.144 uV, v_off_fin = 209.315 uV 18:45:16 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 188.942 uVs 18:45:16 Integrated Voltage Measurement... 18:45:25 Voltage Offset Correction: v_off_ini = 204.932 uV, v_off_fin = 205.110 uV 18:45:25 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -191.125 uVs 18:45:25 Integrated Voltage Measurement... 18:45:34 Voltage Offset Correction: v_off_ini = 221.735 uV, v_off_fin = 226.760 uV 18:45:34 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 166.981 uVs 18:45:34 Integrated Voltage Measurement... 18:45:43 Voltage Offset Correction: v_off_ini = 215.415 uV, v_off_fin = 197.018 uV 18:45:43 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -180.832 uVs 18:45:43 Integrated Voltage Measurement... 18:45:52 Voltage Offset Correction: v_off_ini = 211.711 uV, v_off_fin = 218.331 uV 18:45:52 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 179.858 uVs 18:45:56 I1Y Average and RMS of Measurements VT = -182.134 +- 3.484 uVs I1Y = -36.427 +- 0.697 uTm 18:46:01 Wire Position: X0 = 2.0000 mm, Y0 = 0.0000 mm 18:46:04 Integrated Voltage Measurement... 18:46:13 Voltage Offset Correction: v_off_ini = 198.281 uV, v_off_fin = 196.627 uV 18:46:13 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -110.065 uVs 18:46:13 Integrated Voltage Measurement... 18:46:22 Voltage Offset Correction: v_off_ini = 204.700 uV, v_off_fin = 210.849 uV 18:46:22 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 108.655 uVs 18:46:22 Integrated Voltage Measurement... 18:46:31 Voltage Offset Correction: v_off_ini = 213.711 uV, v_off_fin = 207.295 uV 18:46:31 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -103.146 uVs 18:46:31 Integrated Voltage Measurement... 18:46:40 Voltage Offset Correction: v_off_ini = 221.623 uV, v_off_fin = 211.931 uV 18:46:40 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 112.008 uVs 18:46:40 Integrated Voltage Measurement... 18:46:49 Voltage Offset Correction: v_off_ini = 214.864 uV, v_off_fin = 203.417 uV 18:46:49 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -113.631 uVs 18:46:49 Integrated Voltage Measurement... 18:46:58 Voltage Offset Correction: v_off_ini = 216.462 uV, v_off_fin = 219.029 uV 18:46:58 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 80.065 uVs 18:47:00 I2X Average and RMS of Measurements VT = -104.595 +- 5.526 uVs I2X = 9.623 +- 0.508 uTm^2 18:47:06 Wire Position: X0 = 2.0000 mm, Y0 = 0.0000 mm 18:47:08 Integrated Voltage Measurement... 18:47:17 Voltage Offset Correction: v_off_ini = 198.725 uV, v_off_fin = 192.799 uV 18:47:17 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -268.736 uVs 18:47:17 Integrated Voltage Measurement... 18:47:26 Voltage Offset Correction: v_off_ini = 211.716 uV, v_off_fin = 199.612 uV 18:47:26 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 266.943 uVs 18:47:26 Integrated Voltage Measurement... 18:47:35 Voltage Offset Correction: v_off_ini = 203.470 uV, v_off_fin = 205.166 uV 18:47:35 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -272.307 uVs 18:47:35 Integrated Voltage Measurement... 18:47:44 Voltage Offset Correction: v_off_ini = 201.584 uV, v_off_fin = 202.461 uV 18:47:44 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 276.616 uVs 18:47:44 Integrated Voltage Measurement... 18:47:53 Voltage Offset Correction: v_off_ini = 196.308 uV, v_off_fin = 196.072 uV 18:47:53 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -272.967 uVs 18:47:53 Integrated Voltage Measurement... 18:48:02 Voltage Offset Correction: v_off_ini = 206.296 uV, v_off_fin = 215.492 uV 18:48:02 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 256.230 uVs 18:48:05 I2Y Average and RMS of Measurements VT = -268.966 +- 4.105 uVs I2Y = -24.745 +- 0.378 uTm^2 18:48:13 Wire Position: X0 = 0.0000 mm, Y0 = -1.0000 mm 18:48:17 Integrated Voltage Measurement... 18:48:26 Voltage Offset Correction: v_off_ini = 217.523 uV, v_off_fin = 203.486 uV 18:48:26 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -79.571 uVs 18:48:26 Integrated Voltage Measurement... 18:48:35 Voltage Offset Correction: v_off_ini = 206.445 uV, v_off_fin = 194.700 uV 18:48:35 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 109.352 uVs 18:48:35 Integrated Voltage Measurement... 18:48:44 Voltage Offset Correction: v_off_ini = 208.526 uV, v_off_fin = 213.363 uV 18:48:44 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -96.024 uVs 18:48:44 Integrated Voltage Measurement... 18:48:53 Voltage Offset Correction: v_off_ini = 209.110 uV, v_off_fin = 201.879 uV 18:48:53 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 73.451 uVs 18:48:53 Integrated Voltage Measurement... 18:49:02 Voltage Offset Correction: v_off_ini = 188.343 uV, v_off_fin = 180.896 uV 18:49:02 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -71.940 uVs 18:49:02 Integrated Voltage Measurement... 18:49:11 Voltage Offset Correction: v_off_ini = 193.104 uV, v_off_fin = 194.550 uV 18:49:11 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 84.869 uVs 18:49:15 I1X Average and RMS of Measurements VT = -85.868 +- 6.604 uVs I1X = 17.174 +- 1.321 uTm 18:49:18 Wire Position: X0 = 0.0000 mm, Y0 = -1.0000 mm 18:49:22 Integrated Voltage Measurement... 18:49:31 Voltage Offset Correction: v_off_ini = 187.759 uV, v_off_fin = 172.456 uV 18:49:31 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -182.090 uVs 18:49:31 Integrated Voltage Measurement... 18:49:40 Voltage Offset Correction: v_off_ini = 184.547 uV, v_off_fin = 187.455 uV 18:49:40 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 193.014 uVs 18:49:40 Integrated Voltage Measurement... 18:49:49 Voltage Offset Correction: v_off_ini = 167.639 uV, v_off_fin = 164.833 uV 18:49:49 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -167.639 uVs 18:49:49 Integrated Voltage Measurement... 18:49:58 Voltage Offset Correction: v_off_ini = 189.484 uV, v_off_fin = 202.092 uV 18:49:58 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 176.069 uVs 18:49:58 Integrated Voltage Measurement... 18:50:07 Voltage Offset Correction: v_off_ini = 186.893 uV, v_off_fin = 188.057 uV 18:50:07 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -178.944 uVs 18:50:07 Integrated Voltage Measurement... 18:50:16 Voltage Offset Correction: v_off_ini = 185.436 uV, v_off_fin = 198.504 uV 18:50:16 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 181.333 uVs 18:50:20 I1Y Average and RMS of Measurements VT = -179.848 +- 6.412 uVs I1Y = -35.970 +- 1.282 uTm 18:50:23 Wire Position: X0 = 0.0000 mm, Y0 = -1.0000 mm 18:50:25 Integrated Voltage Measurement... 18:50:34 Voltage Offset Correction: v_off_ini = 172.854 uV, v_off_fin = 175.892 uV 18:50:34 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -94.430 uVs 18:50:34 Integrated Voltage Measurement... 18:50:43 Voltage Offset Correction: v_off_ini = 186.651 uV, v_off_fin = 194.347 uV 18:50:43 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 96.955 uVs 18:50:43 Integrated Voltage Measurement... 18:50:52 Voltage Offset Correction: v_off_ini = 187.146 uV, v_off_fin = 172.607 uV 18:50:52 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -91.826 uVs 18:50:52 Integrated Voltage Measurement... 18:51:01 Voltage Offset Correction: v_off_ini = 180.322 uV, v_off_fin = 188.410 uV 18:51:01 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 94.789 uVs 18:51:01 Integrated Voltage Measurement... 18:51:10 Voltage Offset Correction: v_off_ini = 168.321 uV, v_off_fin = 163.167 uV 18:51:10 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -107.165 uVs 18:51:10 Integrated Voltage Measurement... 18:51:19 Voltage Offset Correction: v_off_ini = 184.937 uV, v_off_fin = 180.256 uV 18:51:19 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 86.509 uVs 18:51:22 I2X Average and RMS of Measurements VT = -95.279 +- 1.470 uVs I2X = 8.766 +- 0.135 uTm^2 18:51:24 Wire Position: X0 = 0.0000 mm, Y0 = -1.0000 mm 18:51:27 Integrated Voltage Measurement... 18:51:36 Voltage Offset Correction: v_off_ini = 159.559 uV, v_off_fin = 154.574 uV 18:51:36 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = -284.269 uVs 18:51:36 Integrated Voltage Measurement... 18:51:45 Voltage Offset Correction: v_off_ini = 153.565 uV, v_off_fin = 165.834 uV 18:51:45 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 295.712 uVs 18:51:45 Integrated Voltage Measurement... 18:51:54 Voltage Offset Correction: v_off_ini = 160.599 uV, v_off_fin = 157.700 uV 18:51:54 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -302.642 uVs 18:51:54 Integrated Voltage Measurement... 18:52:03 Voltage Offset Correction: v_off_ini = 156.452 uV, v_off_fin = 159.975 uV 18:52:03 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 261.389 uVs 18:52:03 Integrated Voltage Measurement... 18:52:12 Voltage Offset Correction: v_off_ini = 156.694 uV, v_off_fin = 159.678 uV 18:52:12 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -301.780 uVs 18:52:12 Integrated Voltage Measurement... 18:52:21 Voltage Offset Correction: v_off_ini = 152.259 uV, v_off_fin = 157.932 uV 18:52:21 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 301.303 uVs 18:52:24 I2Y Average and RMS of Measurements VT = -291.182 +- 8.016 uVs I2Y = -26.789 +- 0.737 uTm^2 18:52:28 Wire Position: X0 = 0.0000 mm, Y0 = -0.5000 mm 18:52:32 Integrated Voltage Measurement... 18:52:41 Voltage Offset Correction: v_off_ini = 156.670 uV, v_off_fin = 153.616 uV 18:52:41 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -74.950 uVs 18:52:41 Integrated Voltage Measurement... 18:52:50 Voltage Offset Correction: v_off_ini = 164.149 uV, v_off_fin = 155.454 uV 18:52:50 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 90.235 uVs 18:52:50 Integrated Voltage Measurement... 18:52:59 Voltage Offset Correction: v_off_ini = 160.487 uV, v_off_fin = 155.198 uV 18:52:59 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -95.570 uVs 18:52:59 Integrated Voltage Measurement... 18:53:08 Voltage Offset Correction: v_off_ini = 171.364 uV, v_off_fin = 162.130 uV 18:53:08 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 112.688 uVs 18:53:08 Integrated Voltage Measurement... 18:53:17 Voltage Offset Correction: v_off_ini = 150.466 uV, v_off_fin = 147.240 uV 18:53:17 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -107.492 uVs 18:53:17 Integrated Voltage Measurement... 18:53:26 Voltage Offset Correction: v_off_ini = 154.787 uV, v_off_fin = 156.012 uV 18:53:26 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 93.996 uVs 18:53:30 I1X Average and RMS of Measurements VT = -95.822 +- 9.456 uVs I1X = 19.165 +- 1.891 uTm 18:53:33 Wire Position: X0 = 0.0000 mm, Y0 = -0.5000 mm 18:53:37 Integrated Voltage Measurement... 18:53:46 Voltage Offset Correction: v_off_ini = 147.154 uV, v_off_fin = 135.619 uV 18:53:46 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -192.068 uVs 18:53:46 Integrated Voltage Measurement... 18:53:55 Voltage Offset Correction: v_off_ini = 147.424 uV, v_off_fin = 157.105 uV 18:53:55 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 188.013 uVs 18:53:55 Integrated Voltage Measurement... 18:54:04 Voltage Offset Correction: v_off_ini = 153.268 uV, v_off_fin = 139.956 uV 18:54:04 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -196.089 uVs 18:54:04 Integrated Voltage Measurement... 18:54:13 Voltage Offset Correction: v_off_ini = 148.277 uV, v_off_fin = 155.410 uV 18:54:13 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 184.798 uVs 18:54:13 Integrated Voltage Measurement... 18:54:22 Voltage Offset Correction: v_off_ini = 148.135 uV, v_off_fin = 135.452 uV 18:54:22 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -193.897 uVs 18:54:22 Integrated Voltage Measurement... 18:54:31 Voltage Offset Correction: v_off_ini = 155.732 uV, v_off_fin = 155.999 uV 18:54:31 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 186.605 uVs 18:54:35 I1Y Average and RMS of Measurements VT = -190.245 +- 0.165 uVs I1Y = -38.049 +- 0.033 uTm 18:54:38 Wire Position: X0 = 0.0000 mm, Y0 = -0.5000 mm 18:54:40 Integrated Voltage Measurement... 18:54:49 Voltage Offset Correction: v_off_ini = 143.226 uV, v_off_fin = 140.798 uV 18:54:49 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -102.769 uVs 18:54:49 Integrated Voltage Measurement... 18:54:58 Voltage Offset Correction: v_off_ini = 144.480 uV, v_off_fin = 155.591 uV 18:54:58 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 115.303 uVs 18:54:59 Integrated Voltage Measurement... 18:55:07 Voltage Offset Correction: v_off_ini = 154.576 uV, v_off_fin = 134.570 uV 18:55:08 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -107.301 uVs 18:55:08 Integrated Voltage Measurement... 18:55:17 Voltage Offset Correction: v_off_ini = 159.033 uV, v_off_fin = 165.351 uV 18:55:17 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 107.567 uVs 18:55:17 Integrated Voltage Measurement... 18:55:26 Voltage Offset Correction: v_off_ini = 160.721 uV, v_off_fin = 152.865 uV 18:55:26 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -101.144 uVs 18:55:26 Integrated Voltage Measurement... 18:55:35 Voltage Offset Correction: v_off_ini = 160.992 uV, v_off_fin = 163.473 uV 18:55:35 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 112.257 uVs 18:55:37 I2X Average and RMS of Measurements VT = -107.724 +- 0.975 uVs I2X = 9.911 +- 0.090 uTm^2 18:55:40 Wire Position: X0 = 0.0000 mm, Y0 = -0.5000 mm 18:55:42 Integrated Voltage Measurement... 18:55:51 Voltage Offset Correction: v_off_ini = 157.437 uV, v_off_fin = 141.480 uV 18:55:51 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -290.950 uVs 18:55:51 Integrated Voltage Measurement... 18:56:00 Voltage Offset Correction: v_off_ini = 140.125 uV, v_off_fin = 138.329 uV 18:56:00 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 276.805 uVs 18:56:00 Integrated Voltage Measurement... 18:56:09 Voltage Offset Correction: v_off_ini = 139.950 uV, v_off_fin = 138.412 uV 18:56:09 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -276.350 uVs 18:56:09 Integrated Voltage Measurement... 18:56:18 Voltage Offset Correction: v_off_ini = 130.806 uV, v_off_fin = 149.693 uV 18:56:18 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 281.826 uVs 18:56:18 Integrated Voltage Measurement... 18:56:27 Voltage Offset Correction: v_off_ini = 141.247 uV, v_off_fin = 139.820 uV 18:56:27 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = -279.720 uVs 18:56:27 Integrated Voltage Measurement... 18:56:36 Voltage Offset Correction: v_off_ini = 149.204 uV, v_off_fin = 154.719 uV 18:56:36 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 283.624 uVs 18:56:39 I2Y Average and RMS of Measurements VT = -281.546 +- 1.957 uVs I2Y = -25.903 +- 0.180 uTm^2 18:56:43 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 18:56:47 Integrated Voltage Measurement... 18:56:56 Voltage Offset Correction: v_off_ini = 126.646 uV, v_off_fin = 125.892 uV 18:56:56 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4998 mm dXY = -0.4999 mm VT = -91.503 uVs 18:56:56 Integrated Voltage Measurement... 18:57:05 Voltage Offset Correction: v_off_ini = 140.714 uV, v_off_fin = 145.780 uV 18:57:05 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 93.095 uVs 18:57:05 Integrated Voltage Measurement... 18:57:14 Voltage Offset Correction: v_off_ini = 142.715 uV, v_off_fin = 122.381 uV 18:57:14 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -75.587 uVs 18:57:14 Integrated Voltage Measurement... 18:57:23 Voltage Offset Correction: v_off_ini = 139.484 uV, v_off_fin = 137.757 uV 18:57:23 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 86.451 uVs 18:57:23 Integrated Voltage Measurement... 18:57:32 Voltage Offset Correction: v_off_ini = 135.138 uV, v_off_fin = 135.530 uV 18:57:32 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -81.253 uVs 18:57:32 Integrated Voltage Measurement... 18:57:41 Voltage Offset Correction: v_off_ini = 141.690 uV, v_off_fin = 153.781 uV 18:57:41 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 114.036 uVs 18:57:45 I1X Average and RMS of Measurements VT = -90.321 +- 6.930 uVs I1X = 18.065 +- 1.386 uTm 18:57:48 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 18:57:52 Integrated Voltage Measurement... 18:58:01 Voltage Offset Correction: v_off_ini = 132.724 uV, v_off_fin = 114.606 uV 18:58:01 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -186.549 uVs 18:58:01 Integrated Voltage Measurement... 18:58:10 Voltage Offset Correction: v_off_ini = 131.897 uV, v_off_fin = 134.981 uV 18:58:10 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 200.097 uVs 18:58:10 Integrated Voltage Measurement... 18:58:19 Voltage Offset Correction: v_off_ini = 136.980 uV, v_off_fin = 130.593 uV 18:58:19 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -185.594 uVs 18:58:19 Integrated Voltage Measurement... 18:58:28 Voltage Offset Correction: v_off_ini = 136.522 uV, v_off_fin = 141.848 uV 18:58:28 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 175.890 uVs 18:58:28 Integrated Voltage Measurement... 18:58:37 Voltage Offset Correction: v_off_ini = 142.640 uV, v_off_fin = 109.840 uV 18:58:37 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -201.253 uVs 18:58:37 Integrated Voltage Measurement... 18:58:46 Voltage Offset Correction: v_off_ini = 128.004 uV, v_off_fin = 127.621 uV 18:58:46 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 196.147 uVs 18:58:50 I1Y Average and RMS of Measurements VT = -190.921 +- 7.525 uVs I1Y = -38.184 +- 1.505 uTm 18:58:53 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 18:58:56 Integrated Voltage Measurement... 18:59:05 Voltage Offset Correction: v_off_ini = 111.711 uV, v_off_fin = 102.236 uV 18:59:05 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -112.041 uVs 18:59:05 Integrated Voltage Measurement... 18:59:14 Voltage Offset Correction: v_off_ini = 110.067 uV, v_off_fin = 114.951 uV 18:59:14 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 110.155 uVs 18:59:14 Integrated Voltage Measurement... 18:59:23 Voltage Offset Correction: v_off_ini = 112.374 uV, v_off_fin = 107.727 uV 18:59:23 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -126.252 uVs 18:59:23 Integrated Voltage Measurement... 18:59:32 Voltage Offset Correction: v_off_ini = 123.258 uV, v_off_fin = 128.151 uV 18:59:32 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 101.328 uVs 18:59:32 Integrated Voltage Measurement... 18:59:41 Voltage Offset Correction: v_off_ini = 112.238 uV, v_off_fin = 107.057 uV 18:59:41 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -113.511 uVs 18:59:41 Integrated Voltage Measurement... 18:59:50 Voltage Offset Correction: v_off_ini = 100.235 uV, v_off_fin = 113.215 uV 18:59:50 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 110.062 uVs 18:59:52 I2X Average and RMS of Measurements VT = -112.225 +- 1.142 uVs I2X = 10.325 +- 0.105 uTm^2 18:59:55 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 18:59:57 Integrated Voltage Measurement... 19:00:06 Voltage Offset Correction: v_off_ini = 97.308 uV, v_off_fin = 106.590 uV 19:00:06 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -297.981 uVs 19:00:06 Integrated Voltage Measurement... 19:00:15 Voltage Offset Correction: v_off_ini = 109.442 uV, v_off_fin = 113.953 uV 19:00:15 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 292.498 uVs 19:00:15 Integrated Voltage Measurement... 19:00:24 Voltage Offset Correction: v_off_ini = 95.852 uV, v_off_fin = 100.066 uV 19:00:24 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -282.959 uVs 19:00:24 Integrated Voltage Measurement... 19:00:33 Voltage Offset Correction: v_off_ini = 99.632 uV, v_off_fin = 99.938 uV 19:00:33 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 281.862 uVs 19:00:33 Integrated Voltage Measurement... 19:00:42 Voltage Offset Correction: v_off_ini = 95.691 uV, v_off_fin = 98.462 uV 19:00:43 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -273.604 uVs 19:00:43 Integrated Voltage Measurement... 19:00:52 Voltage Offset Correction: v_off_ini = 90.677 uV, v_off_fin = 95.532 uV 19:00:52 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 270.463 uVs 19:00:54 I2Y Average and RMS of Measurements VT = -283.228 +- 9.491 uVs I2Y = -26.057 +- 0.873 uTm^2 19:00:58 Wire Position: X0 = 0.0000 mm, Y0 = 0.5000 mm 19:01:02 Integrated Voltage Measurement... 19:01:11 Voltage Offset Correction: v_off_ini = 103.298 uV, v_off_fin = 88.599 uV 19:01:11 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -107.190 uVs 19:01:11 Integrated Voltage Measurement... 19:01:20 Voltage Offset Correction: v_off_ini = 97.353 uV, v_off_fin = 93.617 uV 19:01:20 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 69.686 uVs 19:01:20 Integrated Voltage Measurement... 19:01:29 Voltage Offset Correction: v_off_ini = 100.207 uV, v_off_fin = 80.449 uV 19:01:29 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -92.084 uVs 19:01:29 Integrated Voltage Measurement... 19:01:38 Voltage Offset Correction: v_off_ini = 81.714 uV, v_off_fin = 78.184 uV 19:01:38 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 88.350 uVs 19:01:38 Integrated Voltage Measurement... 19:01:47 Voltage Offset Correction: v_off_ini = 91.286 uV, v_off_fin = 95.516 uV 19:01:47 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -91.242 uVs 19:01:47 Integrated Voltage Measurement... 19:01:56 Voltage Offset Correction: v_off_ini = 94.629 uV, v_off_fin = 90.332 uV 19:01:56 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 93.282 uVs 19:02:01 I1X Average and RMS of Measurements VT = -90.306 +- 1.563 uVs I1X = 18.062 +- 0.312 uTm 19:02:03 Wire Position: X0 = 0.0000 mm, Y0 = 0.5000 mm 19:02:07 Integrated Voltage Measurement... 19:02:16 Voltage Offset Correction: v_off_ini = 102.297 uV, v_off_fin = 92.079 uV 19:02:16 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -188.905 uVs 19:02:16 Integrated Voltage Measurement... 19:02:25 Voltage Offset Correction: v_off_ini = 105.872 uV, v_off_fin = 100.881 uV 19:02:25 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 181.442 uVs 19:02:25 Integrated Voltage Measurement... 19:02:34 Voltage Offset Correction: v_off_ini = 101.527 uV, v_off_fin = 84.982 uV 19:02:34 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -190.487 uVs 19:02:34 Integrated Voltage Measurement... 19:02:43 Voltage Offset Correction: v_off_ini = 98.364 uV, v_off_fin = 110.189 uV 19:02:43 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 187.035 uVs 19:02:43 Integrated Voltage Measurement... 19:02:52 Voltage Offset Correction: v_off_ini = 108.567 uV, v_off_fin = 98.821 uV 19:02:52 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -183.075 uVs 19:02:52 Integrated Voltage Measurement... 19:03:01 Voltage Offset Correction: v_off_ini = 105.117 uV, v_off_fin = 114.650 uV 19:03:01 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 178.855 uVs 19:03:06 I1Y Average and RMS of Measurements VT = -184.967 +- 3.186 uVs I1Y = -36.993 +- 0.637 uTm 19:03:08 Wire Position: X0 = 0.0000 mm, Y0 = 0.5000 mm 19:03:11 Integrated Voltage Measurement... 19:03:20 Voltage Offset Correction: v_off_ini = 118.765 uV, v_off_fin = 106.789 uV 19:03:20 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -93.456 uVs 19:03:20 Integrated Voltage Measurement... 19:03:29 Voltage Offset Correction: v_off_ini = 120.899 uV, v_off_fin = 119.024 uV 19:03:29 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 124.128 uVs 19:03:29 Integrated Voltage Measurement... 19:03:38 Voltage Offset Correction: v_off_ini = 109.315 uV, v_off_fin = 102.990 uV 19:03:38 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -107.257 uVs 19:03:38 Integrated Voltage Measurement... 19:03:47 Voltage Offset Correction: v_off_ini = 116.066 uV, v_off_fin = 127.011 uV 19:03:47 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 100.032 uVs 19:03:47 Integrated Voltage Measurement... 19:03:56 Voltage Offset Correction: v_off_ini = 104.104 uV, v_off_fin = 103.979 uV 19:03:56 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -90.532 uVs 19:03:56 Integrated Voltage Measurement... 19:04:05 Voltage Offset Correction: v_off_ini = 109.554 uV, v_off_fin = 116.730 uV 19:04:05 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 94.572 uVs 19:04:07 I2X Average and RMS of Measurements VT = -101.663 +- 6.776 uVs I2X = 9.353 +- 0.623 uTm^2 19:04:10 Wire Position: X0 = 0.0000 mm, Y0 = 0.5000 mm 19:04:12 Integrated Voltage Measurement... 19:04:21 Voltage Offset Correction: v_off_ini = 106.462 uV, v_off_fin = 96.088 uV 19:04:21 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -288.312 uVs 19:04:22 Integrated Voltage Measurement... 19:04:30 Voltage Offset Correction: v_off_ini = 101.135 uV, v_off_fin = 98.543 uV 19:04:31 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 265.503 uVs 19:04:31 Integrated Voltage Measurement... 19:04:40 Voltage Offset Correction: v_off_ini = 100.170 uV, v_off_fin = 104.683 uV 19:04:40 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -279.979 uVs 19:04:40 Integrated Voltage Measurement... 19:04:49 Voltage Offset Correction: v_off_ini = 95.868 uV, v_off_fin = 98.608 uV 19:04:49 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 278.136 uVs 19:04:49 Integrated Voltage Measurement... 19:04:58 Voltage Offset Correction: v_off_ini = 101.351 uV, v_off_fin = 104.761 uV 19:04:58 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -290.224 uVs 19:04:58 Integrated Voltage Measurement... 19:05:07 Voltage Offset Correction: v_off_ini = 97.851 uV, v_off_fin = 110.498 uV 19:05:07 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 279.685 uVs 19:05:09 I2Y Average and RMS of Measurements VT = -280.307 +- 3.402 uVs I2Y = -25.788 +- 0.313 uTm^2 19:05:13 Wire Position: X0 = 0.0000 mm, Y0 = 1.0000 mm 19:05:17 Integrated Voltage Measurement... 19:05:26 Voltage Offset Correction: v_off_ini = 110.323 uV, v_off_fin = 94.808 uV 19:05:26 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -86.683 uVs 19:05:26 Integrated Voltage Measurement... 19:05:35 Voltage Offset Correction: v_off_ini = 102.068 uV, v_off_fin = 115.722 uV 19:05:35 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 86.981 uVs 19:05:36 Integrated Voltage Measurement... 19:05:44 Voltage Offset Correction: v_off_ini = 120.348 uV, v_off_fin = 97.021 uV 19:05:45 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -79.235 uVs 19:05:45 Integrated Voltage Measurement... 19:05:53 Voltage Offset Correction: v_off_ini = 87.096 uV, v_off_fin = 99.037 uV 19:05:54 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 102.387 uVs 19:05:54 Integrated Voltage Measurement... 19:06:02 Voltage Offset Correction: v_off_ini = 92.739 uV, v_off_fin = 95.287 uV 19:06:03 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -88.228 uVs 19:06:03 Integrated Voltage Measurement... 19:06:11 Voltage Offset Correction: v_off_ini = 88.438 uV, v_off_fin = 97.392 uV 19:06:12 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 96.794 uVs 19:06:16 I1X Average and RMS of Measurements VT = -90.051 +- 2.380 uVs I1X = 18.011 +- 0.476 uTm 19:06:18 Wire Position: X0 = 0.0000 mm, Y0 = 1.0000 mm 19:06:22 Integrated Voltage Measurement... 19:06:31 Voltage Offset Correction: v_off_ini = 89.331 uV, v_off_fin = 86.470 uV 19:06:31 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -182.818 uVs 19:06:31 Integrated Voltage Measurement... 19:06:40 Voltage Offset Correction: v_off_ini = 88.689 uV, v_off_fin = 93.432 uV 19:06:40 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 172.007 uVs 19:06:40 Integrated Voltage Measurement... 19:06:49 Voltage Offset Correction: v_off_ini = 66.106 uV, v_off_fin = 62.470 uV 19:06:49 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -160.739 uVs 19:06:49 Integrated Voltage Measurement... 19:06:58 Voltage Offset Correction: v_off_ini = 77.023 uV, v_off_fin = 93.809 uV 19:06:58 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 184.061 uVs 19:06:58 Integrated Voltage Measurement... 19:07:07 Voltage Offset Correction: v_off_ini = 88.093 uV, v_off_fin = 72.190 uV 19:07:07 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -171.992 uVs 19:07:07 Integrated Voltage Measurement... 19:07:16 Voltage Offset Correction: v_off_ini = 81.065 uV, v_off_fin = 91.541 uV 19:07:16 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 192.703 uVs 19:07:21 I1Y Average and RMS of Measurements VT = -177.387 +- 4.061 uVs I1Y = -35.474 +- 0.812 uTm 19:07:23 Wire Position: X0 = 0.0000 mm, Y0 = 1.0000 mm 19:07:26 Integrated Voltage Measurement... 19:07:35 Voltage Offset Correction: v_off_ini = 85.664 uV, v_off_fin = 78.012 uV 19:07:35 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -81.929 uVs 19:07:35 Integrated Voltage Measurement... 19:07:44 Voltage Offset Correction: v_off_ini = 77.746 uV, v_off_fin = 88.080 uV 19:07:44 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 104.999 uVs 19:07:44 Integrated Voltage Measurement... 19:07:53 Voltage Offset Correction: v_off_ini = 85.165 uV, v_off_fin = 79.815 uV 19:07:53 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -109.639 uVs 19:07:53 Integrated Voltage Measurement... 19:08:02 Voltage Offset Correction: v_off_ini = 78.804 uV, v_off_fin = 96.895 uV 19:08:02 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 106.355 uVs 19:08:02 Integrated Voltage Measurement... 19:08:11 Voltage Offset Correction: v_off_ini = 87.814 uV, v_off_fin = 86.646 uV 19:08:11 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -97.140 uVs 19:08:11 Integrated Voltage Measurement... 19:08:20 Voltage Offset Correction: v_off_ini = 88.236 uV, v_off_fin = 100.945 uV 19:08:20 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 100.073 uVs 19:08:22 I2X Average and RMS of Measurements VT = -100.022 +- 6.017 uVs I2X = 9.202 +- 0.554 uTm^2 19:08:25 Wire Position: X0 = 0.0000 mm, Y0 = 1.0000 mm 19:08:27 Integrated Voltage Measurement... 19:08:36 Voltage Offset Correction: v_off_ini = 92.834 uV, v_off_fin = 85.817 uV 19:08:36 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -290.718 uVs 19:08:36 Integrated Voltage Measurement... 19:08:45 Voltage Offset Correction: v_off_ini = 82.894 uV, v_off_fin = 87.885 uV 19:08:45 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 287.649 uVs 19:08:45 Integrated Voltage Measurement... 19:08:54 Voltage Offset Correction: v_off_ini = 96.112 uV, v_off_fin = 83.836 uV 19:08:54 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -283.387 uVs 19:08:54 Integrated Voltage Measurement... 19:09:03 Voltage Offset Correction: v_off_ini = 94.031 uV, v_off_fin = 87.371 uV 19:09:03 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 295.435 uVs 19:09:03 Integrated Voltage Measurement... 19:09:12 Voltage Offset Correction: v_off_ini = 82.985 uV, v_off_fin = 94.839 uV 19:09:12 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -265.702 uVs 19:09:12 Integrated Voltage Measurement... 19:09:21 Voltage Offset Correction: v_off_ini = 92.775 uV, v_off_fin = 100.591 uV 19:09:21 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 282.945 uVs 19:09:24 I2Y Average and RMS of Measurements VT = -284.306 +- 7.059 uVs I2Y = -26.156 +- 0.649 uTm^2 19:09:41 Wire Position: X0 = -2.0000 mm, Y0 = 0.0000 mm 19:09:45 Integrated Voltage Measurement... 19:09:54 Voltage Offset Correction: v_off_ini = 104.741 uV, v_off_fin = 92.106 uV 19:09:54 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -112.094 uVs 19:09:54 Integrated Voltage Measurement... 19:10:03 Voltage Offset Correction: v_off_ini = 105.886 uV, v_off_fin = 117.658 uV 19:10:03 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 109.101 uVs 19:10:03 Integrated Voltage Measurement... 19:10:12 Voltage Offset Correction: v_off_ini = 106.208 uV, v_off_fin = 98.494 uV 19:10:12 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -106.282 uVs 19:10:12 Integrated Voltage Measurement... 19:10:21 Voltage Offset Correction: v_off_ini = 94.452 uV, v_off_fin = 100.473 uV 19:10:21 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 117.395 uVs 19:10:21 Integrated Voltage Measurement... 19:10:30 Voltage Offset Correction: v_off_ini = 100.596 uV, v_off_fin = 104.243 uV 19:10:30 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -99.617 uVs 19:10:30 Integrated Voltage Measurement... 19:10:39 Voltage Offset Correction: v_off_ini = 113.248 uV, v_off_fin = 113.592 uV 19:10:39 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 102.626 uVs 19:10:43 I1X Average and RMS of Measurements VT = -107.852 +- 4.786 uVs I1X = 21.571 +- 0.958 uTm 19:10:46 Wire Position: X0 = -2.0000 mm, Y0 = 0.0000 mm 19:10:50 Integrated Voltage Measurement... 19:10:59 Voltage Offset Correction: v_off_ini = 109.272 uV, v_off_fin = 105.806 uV 19:10:59 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -187.778 uVs 19:10:59 Integrated Voltage Measurement... 19:11:08 Voltage Offset Correction: v_off_ini = 109.896 uV, v_off_fin = 124.740 uV 19:11:08 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 183.929 uVs 19:11:08 Integrated Voltage Measurement... 19:11:17 Voltage Offset Correction: v_off_ini = 120.205 uV, v_off_fin = 107.996 uV 19:11:17 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -195.639 uVs 19:11:17 Integrated Voltage Measurement... 19:11:26 Voltage Offset Correction: v_off_ini = 116.781 uV, v_off_fin = 125.355 uV 19:11:26 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 183.062 uVs 19:11:26 Integrated Voltage Measurement... 19:11:35 Voltage Offset Correction: v_off_ini = 119.431 uV, v_off_fin = 118.660 uV 19:11:35 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -186.162 uVs 19:11:35 Integrated Voltage Measurement... 19:11:44 Voltage Offset Correction: v_off_ini = 124.637 uV, v_off_fin = 127.122 uV 19:11:44 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 189.330 uVs 19:11:48 I1Y Average and RMS of Measurements VT = -187.650 +- 1.429 uVs I1Y = -37.530 +- 0.286 uTm 19:11:53 Wire Position: X0 = -2.0000 mm, Y0 = 0.0000 mm 19:11:56 Integrated Voltage Measurement... 19:12:05 Voltage Offset Correction: v_off_ini = 111.440 uV, v_off_fin = 114.344 uV 19:12:05 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -100.184 uVs 19:12:05 Integrated Voltage Measurement... 19:12:14 Voltage Offset Correction: v_off_ini = 140.813 uV, v_off_fin = 133.063 uV 19:12:14 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 111.148 uVs 19:12:14 Integrated Voltage Measurement... 19:12:23 Voltage Offset Correction: v_off_ini = 114.995 uV, v_off_fin = 122.174 uV 19:12:23 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -127.504 uVs 19:12:23 Integrated Voltage Measurement... 19:12:32 Voltage Offset Correction: v_off_ini = 126.645 uV, v_off_fin = 128.804 uV 19:12:32 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 118.922 uVs 19:12:32 Integrated Voltage Measurement... 19:12:41 Voltage Offset Correction: v_off_ini = 124.842 uV, v_off_fin = 112.892 uV 19:12:41 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -99.565 uVs 19:12:41 Integrated Voltage Measurement... 19:12:50 Voltage Offset Correction: v_off_ini = 114.473 uV, v_off_fin = 120.694 uV 19:12:50 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 133.521 uVs 19:12:52 I2X Average and RMS of Measurements VT = -115.141 +- 7.232 uVs I2X = 10.593 +- 0.665 uTm^2 19:12:57 Wire Position: X0 = -2.0000 mm, Y0 = 0.0000 mm 19:13:00 Integrated Voltage Measurement... 19:13:09 Voltage Offset Correction: v_off_ini = 125.135 uV, v_off_fin = 115.892 uV 19:13:09 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -292.464 uVs 19:13:09 Integrated Voltage Measurement... 19:13:18 Voltage Offset Correction: v_off_ini = 112.131 uV, v_off_fin = 114.835 uV 19:13:18 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 283.125 uVs 19:13:18 Integrated Voltage Measurement... 19:13:27 Voltage Offset Correction: v_off_ini = 105.590 uV, v_off_fin = 101.629 uV 19:13:27 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -285.929 uVs 19:13:27 Integrated Voltage Measurement... 19:13:36 Voltage Offset Correction: v_off_ini = 100.464 uV, v_off_fin = 97.490 uV 19:13:36 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 272.980 uVs 19:13:36 Integrated Voltage Measurement... 19:13:45 Voltage Offset Correction: v_off_ini = 102.167 uV, v_off_fin = 100.587 uV 19:13:45 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -268.457 uVs 19:13:45 Integrated Voltage Measurement... 19:13:54 Voltage Offset Correction: v_off_ini = 86.574 uV, v_off_fin = 101.206 uV 19:13:54 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 283.581 uVs 19:13:57 I2Y Average and RMS of Measurements VT = -281.089 +- 4.944 uVs I2Y = -25.860 +- 0.455 uTm^2 19:14:01 Wire Position: X0 = -1.0000 mm, Y0 = 0.0000 mm 19:14:06 Integrated Voltage Measurement... 19:14:14 Voltage Offset Correction: v_off_ini = 85.832 uV, v_off_fin = 82.333 uV 19:14:14 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -97.206 uVs 19:14:15 Integrated Voltage Measurement... 19:14:23 Voltage Offset Correction: v_off_ini = 84.595 uV, v_off_fin = 84.004 uV 19:14:24 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 112.702 uVs 19:14:24 Integrated Voltage Measurement... 19:14:32 Voltage Offset Correction: v_off_ini = 84.178 uV, v_off_fin = 82.161 uV 19:14:33 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -98.993 uVs 19:14:33 Integrated Voltage Measurement... 19:14:41 Voltage Offset Correction: v_off_ini = 75.258 uV, v_off_fin = 83.663 uV 19:14:42 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 109.764 uVs 19:14:42 Integrated Voltage Measurement... 19:14:51 Voltage Offset Correction: v_off_ini = 77.422 uV, v_off_fin = 74.846 uV 19:14:51 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -119.567 uVs 19:14:51 Integrated Voltage Measurement... 19:15:00 Voltage Offset Correction: v_off_ini = 90.689 uV, v_off_fin = 86.294 uV 19:15:00 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 109.158 uVs 19:15:04 I1X Average and RMS of Measurements VT = -107.898 +- 4.577 uVs I1X = 21.580 +- 0.915 uTm 19:15:06 Wire Position: X0 = -1.0000 mm, Y0 = 0.0000 mm 19:15:11 Integrated Voltage Measurement... 19:15:19 Voltage Offset Correction: v_off_ini = 86.377 uV, v_off_fin = 76.267 uV 19:15:19 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -182.637 uVs 19:15:20 Integrated Voltage Measurement... 19:15:28 Voltage Offset Correction: v_off_ini = 92.106 uV, v_off_fin = 89.665 uV 19:15:28 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 182.933 uVs 19:15:28 Integrated Voltage Measurement... 19:15:37 Voltage Offset Correction: v_off_ini = 95.055 uV, v_off_fin = 80.540 uV 19:15:37 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -195.318 uVs 19:15:37 Integrated Voltage Measurement... 19:15:46 Voltage Offset Correction: v_off_ini = 91.242 uV, v_off_fin = 90.551 uV 19:15:46 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 184.666 uVs 19:15:46 Integrated Voltage Measurement... 19:15:55 Voltage Offset Correction: v_off_ini = 85.027 uV, v_off_fin = 87.590 uV 19:15:55 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -179.086 uVs 19:15:55 Integrated Voltage Measurement... 19:16:04 Voltage Offset Correction: v_off_ini = 95.492 uV, v_off_fin = 103.595 uV 19:16:04 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 178.560 uVs 19:16:09 I1Y Average and RMS of Measurements VT = -183.867 +- 4.624 uVs I1Y = -36.773 +- 0.925 uTm 19:16:14 Wire Position: X0 = -1.0000 mm, Y0 = 0.0000 mm 19:16:16 Integrated Voltage Measurement... 19:16:25 Voltage Offset Correction: v_off_ini = 83.021 uV, v_off_fin = 77.369 uV 19:16:25 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -136.110 uVs 19:16:25 Integrated Voltage Measurement... 19:16:34 Voltage Offset Correction: v_off_ini = 93.296 uV, v_off_fin = 89.788 uV 19:16:34 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 119.615 uVs 19:16:34 Integrated Voltage Measurement... 19:16:43 Voltage Offset Correction: v_off_ini = 81.770 uV, v_off_fin = 73.227 uV 19:16:43 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -135.093 uVs 19:16:43 Integrated Voltage Measurement... 19:16:52 Voltage Offset Correction: v_off_ini = 92.681 uV, v_off_fin = 102.587 uV 19:16:52 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 114.980 uVs 19:16:52 Integrated Voltage Measurement... 19:17:01 Voltage Offset Correction: v_off_ini = 102.836 uV, v_off_fin = 79.314 uV 19:17:01 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -122.287 uVs 19:17:01 Integrated Voltage Measurement... 19:17:10 Voltage Offset Correction: v_off_ini = 95.487 uV, v_off_fin = 104.818 uV 19:17:10 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 126.330 uVs 19:17:13 I2X Average and RMS of Measurements VT = -125.736 +- 1.533 uVs I2X = 11.568 +- 0.141 uTm^2 19:17:18 Wire Position: X0 = -1.0000 mm, Y0 = 0.0000 mm 19:17:21 Integrated Voltage Measurement... 19:17:29 Voltage Offset Correction: v_off_ini = 100.227 uV, v_off_fin = 92.796 uV 19:17:30 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -275.441 uVs 19:17:30 Integrated Voltage Measurement... 19:17:39 Voltage Offset Correction: v_off_ini = 102.652 uV, v_off_fin = 98.937 uV 19:17:39 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 271.725 uVs 19:17:39 Integrated Voltage Measurement... 19:17:48 Voltage Offset Correction: v_off_ini = 110.541 uV, v_off_fin = 97.768 uV 19:17:48 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -291.230 uVs 19:17:48 Integrated Voltage Measurement... 19:17:57 Voltage Offset Correction: v_off_ini = 84.614 uV, v_off_fin = 96.572 uV 19:17:57 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 264.849 uVs 19:17:57 Integrated Voltage Measurement... 19:18:06 Voltage Offset Correction: v_off_ini = 87.339 uV, v_off_fin = 97.224 uV 19:18:06 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -269.095 uVs 19:18:06 Integrated Voltage Measurement... 19:18:15 Voltage Offset Correction: v_off_ini = 93.216 uV, v_off_fin = 90.131 uV 19:18:15 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 258.900 uVs 19:18:17 I2Y Average and RMS of Measurements VT = -271.873 +- 5.858 uVs I2Y = -25.012 +- 0.539 uTm^2 19:18:22 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 19:18:26 Integrated Voltage Measurement... 19:18:35 Voltage Offset Correction: v_off_ini = 96.088 uV, v_off_fin = 84.583 uV 19:18:35 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -77.575 uVs 19:18:35 Integrated Voltage Measurement... 19:18:44 Voltage Offset Correction: v_off_ini = 93.074 uV, v_off_fin = 102.579 uV 19:18:44 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 85.622 uVs 19:18:44 Integrated Voltage Measurement... 19:18:53 Voltage Offset Correction: v_off_ini = 92.704 uV, v_off_fin = 96.899 uV 19:18:53 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -103.018 uVs 19:18:53 Integrated Voltage Measurement... 19:19:02 Voltage Offset Correction: v_off_ini = 114.556 uV, v_off_fin = 115.718 uV 19:19:02 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = 104.506 uVs 19:19:02 Integrated Voltage Measurement... 19:19:11 Voltage Offset Correction: v_off_ini = 110.166 uV, v_off_fin = 94.404 uV 19:19:11 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -100.093 uVs 19:19:11 Integrated Voltage Measurement... 19:19:20 Voltage Offset Correction: v_off_ini = 84.137 uV, v_off_fin = 96.420 uV 19:19:20 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 118.219 uVs 19:19:24 I1X Average and RMS of Measurements VT = -98.172 +- 11.925 uVs I1X = 19.635 +- 2.385 uTm 19:19:27 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 19:19:31 Integrated Voltage Measurement... 19:19:40 Voltage Offset Correction: v_off_ini = 85.210 uV, v_off_fin = 85.196 uV 19:19:40 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -173.071 uVs 19:19:40 Integrated Voltage Measurement... 19:19:49 Voltage Offset Correction: v_off_ini = 100.926 uV, v_off_fin = 105.997 uV 19:19:49 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 178.067 uVs 19:19:49 Integrated Voltage Measurement... 19:19:58 Voltage Offset Correction: v_off_ini = 96.657 uV, v_off_fin = 100.726 uV 19:19:58 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -198.132 uVs 19:19:58 Integrated Voltage Measurement... 19:20:07 Voltage Offset Correction: v_off_ini = 92.683 uV, v_off_fin = 102.431 uV 19:20:07 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 200.078 uVs 19:20:07 Integrated Voltage Measurement... 19:20:16 Voltage Offset Correction: v_off_ini = 103.880 uV, v_off_fin = 89.950 uV 19:20:16 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -181.633 uVs 19:20:16 Integrated Voltage Measurement... 19:20:25 Voltage Offset Correction: v_off_ini = 102.207 uV, v_off_fin = 114.734 uV 19:20:25 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 164.623 uVs 19:20:29 I1Y Average and RMS of Measurements VT = -182.601 +- 11.713 uVs I1Y = -36.520 +- 2.343 uTm 19:20:34 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 19:20:37 Integrated Voltage Measurement... 19:20:46 Voltage Offset Correction: v_off_ini = 108.170 uV, v_off_fin = 97.843 uV 19:20:46 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -120.085 uVs 19:20:46 Integrated Voltage Measurement... 19:20:55 Voltage Offset Correction: v_off_ini = 97.601 uV, v_off_fin = 107.388 uV 19:20:55 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 145.364 uVs 19:20:55 Integrated Voltage Measurement... 19:21:04 Voltage Offset Correction: v_off_ini = 115.905 uV, v_off_fin = 95.406 uV 19:21:04 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -122.884 uVs 19:21:04 Integrated Voltage Measurement... 19:21:13 Voltage Offset Correction: v_off_ini = 106.239 uV, v_off_fin = 112.839 uV 19:21:13 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 126.335 uVs 19:21:13 Integrated Voltage Measurement... 19:21:22 Voltage Offset Correction: v_off_ini = 95.553 uV, v_off_fin = 96.392 uV 19:21:22 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -113.530 uVs 19:21:22 Integrated Voltage Measurement... 19:21:31 Voltage Offset Correction: v_off_ini = 118.012 uV, v_off_fin = 108.049 uV 19:21:31 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 123.933 uVs 19:21:33 I2X Average and RMS of Measurements VT = -125.355 +- 5.737 uVs I2X = 11.533 +- 0.528 uTm^2 19:21:38 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 19:21:41 Integrated Voltage Measurement... 19:21:50 Voltage Offset Correction: v_off_ini = 112.560 uV, v_off_fin = 96.583 uV 19:21:50 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -272.715 uVs 19:21:50 Integrated Voltage Measurement... 19:21:59 Voltage Offset Correction: v_off_ini = 102.686 uV, v_off_fin = 114.589 uV 19:21:59 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 271.678 uVs 19:21:59 Integrated Voltage Measurement... 19:22:08 Voltage Offset Correction: v_off_ini = 115.482 uV, v_off_fin = 112.430 uV 19:22:08 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -271.216 uVs 19:22:08 Integrated Voltage Measurement... 19:22:17 Voltage Offset Correction: v_off_ini = 109.269 uV, v_off_fin = 108.463 uV 19:22:17 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 267.016 uVs 19:22:17 Integrated Voltage Measurement... 19:22:26 Voltage Offset Correction: v_off_ini = 112.395 uV, v_off_fin = 113.150 uV 19:22:26 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -283.288 uVs 19:22:26 Integrated Voltage Measurement... 19:22:35 Voltage Offset Correction: v_off_ini = 110.484 uV, v_off_fin = 101.430 uV 19:22:35 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 260.405 uVs 19:22:37 I2Y Average and RMS of Measurements VT = -271.053 +- 1.377 uVs I2Y = -24.937 +- 0.127 uTm^2 19:22:42 Wire Position: X0 = 1.0000 mm, Y0 = 0.0000 mm 19:22:46 Integrated Voltage Measurement... 19:22:55 Voltage Offset Correction: v_off_ini = 98.424 uV, v_off_fin = 95.942 uV 19:22:55 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4997 mm dXY = -0.4999 mm VT = -104.334 uVs 19:22:55 Integrated Voltage Measurement... 19:23:04 Voltage Offset Correction: v_off_ini = 105.146 uV, v_off_fin = 108.947 uV 19:23:04 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 105.182 uVs 19:23:04 Integrated Voltage Measurement... 19:23:13 Voltage Offset Correction: v_off_ini = 102.897 uV, v_off_fin = 104.202 uV 19:23:13 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -93.206 uVs 19:23:13 Integrated Voltage Measurement... 19:23:22 Voltage Offset Correction: v_off_ini = 100.815 uV, v_off_fin = 115.050 uV 19:23:22 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 119.066 uVs 19:23:22 Integrated Voltage Measurement... 19:23:31 Voltage Offset Correction: v_off_ini = 100.920 uV, v_off_fin = 95.050 uV 19:23:31 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -103.789 uVs 19:23:31 Integrated Voltage Measurement... 19:23:40 Voltage Offset Correction: v_off_ini = 102.811 uV, v_off_fin = 104.940 uV 19:23:40 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 108.492 uVs 19:23:45 I1X Average and RMS of Measurements VT = -105.678 +- 0.651 uVs I1X = 21.137 +- 0.129 uTm 19:23:47 Wire Position: X0 = 1.0000 mm, Y0 = 0.0000 mm 19:23:51 Integrated Voltage Measurement... 19:24:00 Voltage Offset Correction: v_off_ini = 88.453 uV, v_off_fin = 92.207 uV 19:24:00 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -180.166 uVs 19:24:00 Integrated Voltage Measurement... 19:24:09 Voltage Offset Correction: v_off_ini = 111.729 uV, v_off_fin = 118.182 uV 19:24:09 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 184.391 uVs 19:24:09 Integrated Voltage Measurement... 19:24:18 Voltage Offset Correction: v_off_ini = 107.245 uV, v_off_fin = 97.585 uV 19:24:18 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -184.659 uVs 19:24:18 Integrated Voltage Measurement... 19:24:27 Voltage Offset Correction: v_off_ini = 104.798 uV, v_off_fin = 106.014 uV 19:24:27 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 188.515 uVs 19:24:27 Integrated Voltage Measurement... 19:24:36 Voltage Offset Correction: v_off_ini = 117.736 uV, v_off_fin = 91.360 uV 19:24:36 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -179.039 uVs 19:24:36 Integrated Voltage Measurement... 19:24:45 Voltage Offset Correction: v_off_ini = 100.725 uV, v_off_fin = 101.062 uV 19:24:45 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 184.178 uVs 19:24:50 I1Y Average and RMS of Measurements VT = -183.491 +- 2.206 uVs I1Y = -36.698 +- 0.441 uTm 19:24:55 Wire Position: X0 = 1.0000 mm, Y0 = 0.0000 mm 19:24:57 Integrated Voltage Measurement... 19:25:06 Voltage Offset Correction: v_off_ini = 96.228 uV, v_off_fin = 86.804 uV 19:25:06 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -110.485 uVs 19:25:06 Integrated Voltage Measurement... 19:25:15 Voltage Offset Correction: v_off_ini = 102.712 uV, v_off_fin = 107.103 uV 19:25:15 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 124.036 uVs 19:25:15 Integrated Voltage Measurement... 19:25:24 Voltage Offset Correction: v_off_ini = 101.115 uV, v_off_fin = 96.311 uV 19:25:24 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -108.460 uVs 19:25:25 Integrated Voltage Measurement... 19:25:33 Voltage Offset Correction: v_off_ini = 94.215 uV, v_off_fin = 112.040 uV 19:25:34 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 114.792 uVs 19:25:34 Integrated Voltage Measurement... 19:25:43 Voltage Offset Correction: v_off_ini = 98.499 uV, v_off_fin = 99.560 uV 19:25:43 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -132.813 uVs 19:25:43 Integrated Voltage Measurement... 19:25:52 Voltage Offset Correction: v_off_ini = 105.926 uV, v_off_fin = 117.862 uV 19:25:52 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 130.135 uVs 19:25:54 I2X Average and RMS of Measurements VT = -120.120 +- 8.351 uVs I2X = 11.051 +- 0.768 uTm^2 19:25:59 Wire Position: X0 = 1.0000 mm, Y0 = 0.0000 mm 19:26:02 Integrated Voltage Measurement... 19:26:11 Voltage Offset Correction: v_off_ini = 100.977 uV, v_off_fin = 94.160 uV 19:26:11 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -278.320 uVs 19:26:11 Integrated Voltage Measurement... 19:26:20 Voltage Offset Correction: v_off_ini = 108.557 uV, v_off_fin = 108.901 uV 19:26:20 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 258.111 uVs 19:26:20 Integrated Voltage Measurement... 19:26:29 Voltage Offset Correction: v_off_ini = 99.408 uV, v_off_fin = 96.636 uV 19:26:29 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -263.581 uVs 19:26:29 Integrated Voltage Measurement... 19:26:38 Voltage Offset Correction: v_off_ini = 102.080 uV, v_off_fin = 98.605 uV 19:26:38 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 266.212 uVs 19:26:38 Integrated Voltage Measurement... 19:26:47 Voltage Offset Correction: v_off_ini = 109.032 uV, v_off_fin = 105.674 uV 19:26:47 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -277.555 uVs 19:26:47 Integrated Voltage Measurement... 19:26:56 Voltage Offset Correction: v_off_ini = 93.829 uV, v_off_fin = 107.951 uV 19:26:56 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 264.943 uVs 19:26:58 I2Y Average and RMS of Measurements VT = -268.120 +- 2.594 uVs I2Y = -24.667 +- 0.239 uTm^2 19:27:03 Wire Position: X0 = 2.0000 mm, Y0 = 0.0000 mm 19:27:07 Integrated Voltage Measurement... 19:27:16 Voltage Offset Correction: v_off_ini = 120.858 uV, v_off_fin = 112.166 uV 19:27:16 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.4999 mm dXY = -0.4999 mm VT = -140.180 uVs 19:27:16 Integrated Voltage Measurement... 19:27:25 Voltage Offset Correction: v_off_ini = 99.971 uV, v_off_fin = 119.269 uV 19:27:25 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = 130.523 uVs 19:27:25 Integrated Voltage Measurement... 19:27:34 Voltage Offset Correction: v_off_ini = 118.679 uV, v_off_fin = 115.061 uV 19:27:34 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.4999 mm dXY = -0.4999 mm VT = -118.044 uVs 19:27:34 Integrated Voltage Measurement... 19:27:43 Voltage Offset Correction: v_off_ini = 112.638 uV, v_off_fin = 123.043 uV 19:27:43 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 100.587 uVs 19:27:43 Integrated Voltage Measurement... 19:27:52 Voltage Offset Correction: v_off_ini = 113.040 uV, v_off_fin = 115.920 uV 19:27:52 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -88.843 uVs 19:27:52 Integrated Voltage Measurement... 19:28:01 Voltage Offset Correction: v_off_ini = 103.251 uV, v_off_fin = 102.352 uV 19:28:01 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 101.240 uVs 19:28:06 I1X Average and RMS of Measurements VT = -113.236 +- 16.688 uVs I1X = 22.649 +- 3.339 uTm 19:28:08 Wire Position: X0 = 2.0000 mm, Y0 = 0.0000 mm 19:28:12 Integrated Voltage Measurement... 19:28:21 Voltage Offset Correction: v_off_ini = 101.947 uV, v_off_fin = 104.185 uV 19:28:21 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -179.297 uVs 19:28:21 Integrated Voltage Measurement... 19:28:30 Voltage Offset Correction: v_off_ini = 96.876 uV, v_off_fin = 110.514 uV 19:28:30 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 182.760 uVs 19:28:30 Integrated Voltage Measurement... 19:28:39 Voltage Offset Correction: v_off_ini = 92.044 uV, v_off_fin = 93.784 uV 19:28:39 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -176.708 uVs 19:28:39 Integrated Voltage Measurement... 19:28:48 Voltage Offset Correction: v_off_ini = 94.874 uV, v_off_fin = 110.899 uV 19:28:48 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 185.408 uVs 19:28:48 Integrated Voltage Measurement... 19:28:57 Voltage Offset Correction: v_off_ini = 96.525 uV, v_off_fin = 86.180 uV 19:28:57 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -191.099 uVs 19:28:57 Integrated Voltage Measurement... 19:29:06 Voltage Offset Correction: v_off_ini = 97.245 uV, v_off_fin = 102.990 uV 19:29:06 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 165.366 uVs 19:29:11 I1Y Average and RMS of Measurements VT = -180.106 +- 1.325 uVs I1Y = -36.021 +- 0.265 uTm 19:29:16 Wire Position: X0 = 2.0000 mm, Y0 = 0.0000 mm 19:29:18 Integrated Voltage Measurement... 19:29:27 Voltage Offset Correction: v_off_ini = 94.202 uV, v_off_fin = 87.225 uV 19:29:27 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -104.606 uVs 19:29:27 Integrated Voltage Measurement... 19:29:36 Voltage Offset Correction: v_off_ini = 85.174 uV, v_off_fin = 112.269 uV 19:29:36 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 112.381 uVs 19:29:36 Integrated Voltage Measurement... 19:29:45 Voltage Offset Correction: v_off_ini = 106.853 uV, v_off_fin = 101.809 uV 19:29:45 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -120.915 uVs 19:29:46 Integrated Voltage Measurement... 19:29:55 Voltage Offset Correction: v_off_ini = 104.597 uV, v_off_fin = 122.315 uV 19:29:55 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 118.313 uVs 19:29:55 Integrated Voltage Measurement... 19:30:04 Voltage Offset Correction: v_off_ini = 116.794 uV, v_off_fin = 97.775 uV 19:30:04 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -105.049 uVs 19:30:04 Integrated Voltage Measurement... 19:30:13 Voltage Offset Correction: v_off_ini = 103.403 uV, v_off_fin = 107.523 uV 19:30:13 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 124.160 uVs 19:30:15 I2X Average and RMS of Measurements VT = -114.238 +- 4.547 uVs I2X = 10.510 +- 0.418 uTm^2 19:30:20 Wire Position: X0 = 2.0000 mm, Y0 = 0.0000 mm 19:30:23 Integrated Voltage Measurement... 19:30:32 Voltage Offset Correction: v_off_ini = 106.666 uV, v_off_fin = 102.696 uV 19:30:32 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -250.114 uVs 19:30:32 Integrated Voltage Measurement... 19:30:41 Voltage Offset Correction: v_off_ini = 117.660 uV, v_off_fin = 113.669 uV 19:30:41 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 252.659 uVs 19:30:41 Integrated Voltage Measurement... 19:30:50 Voltage Offset Correction: v_off_ini = 101.843 uV, v_off_fin = 91.306 uV 19:30:50 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -251.250 uVs 19:30:50 Integrated Voltage Measurement... 19:30:59 Voltage Offset Correction: v_off_ini = 92.532 uV, v_off_fin = 99.189 uV 19:30:59 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 273.369 uVs 19:30:59 Integrated Voltage Measurement... 19:31:08 Voltage Offset Correction: v_off_ini = 96.208 uV, v_off_fin = 105.083 uV 19:31:08 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -291.110 uVs 19:31:08 Integrated Voltage Measurement... 19:31:17 Voltage Offset Correction: v_off_ini = 99.280 uV, v_off_fin = 99.447 uV 19:31:17 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 261.870 uVs 19:31:19 I2Y Average and RMS of Measurements VT = -263.395 +- 10.277 uVs I2Y = -24.232 +- 0.946 uTm^2 19:31:27 Wire Position: X0 = 0.0000 mm, Y0 = -1.0000 mm 19:31:31 Integrated Voltage Measurement... 19:31:40 Voltage Offset Correction: v_off_ini = 106.095 uV, v_off_fin = 92.943 uV 19:31:40 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -111.209 uVs 19:31:40 Integrated Voltage Measurement... 19:31:49 Voltage Offset Correction: v_off_ini = 99.741 uV, v_off_fin = 103.913 uV 19:31:49 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 113.101 uVs 19:31:50 Integrated Voltage Measurement... 19:31:58 Voltage Offset Correction: v_off_ini = 96.789 uV, v_off_fin = 117.603 uV 19:31:58 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -82.374 uVs 19:31:59 Integrated Voltage Measurement... 19:32:07 Voltage Offset Correction: v_off_ini = 119.107 uV, v_off_fin = 112.346 uV 19:32:08 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 116.292 uVs 19:32:08 Integrated Voltage Measurement... 19:32:16 Voltage Offset Correction: v_off_ini = 114.574 uV, v_off_fin = 104.786 uV 19:32:17 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -105.316 uVs 19:32:17 Integrated Voltage Measurement... 19:32:26 Voltage Offset Correction: v_off_ini = 103.723 uV, v_off_fin = 91.556 uV 19:32:26 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 94.450 uVs 19:32:30 I1X Average and RMS of Measurements VT = -103.790 +- 5.919 uVs I1X = 20.758 +- 1.184 uTm 19:32:32 Wire Position: X0 = 0.0000 mm, Y0 = -1.0000 mm 19:32:37 Integrated Voltage Measurement... 19:32:45 Voltage Offset Correction: v_off_ini = 99.650 uV, v_off_fin = 89.917 uV 19:32:46 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -182.170 uVs 19:32:46 Integrated Voltage Measurement... 19:32:54 Voltage Offset Correction: v_off_ini = 106.201 uV, v_off_fin = 114.644 uV 19:32:55 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 175.062 uVs 19:32:55 Integrated Voltage Measurement... 19:33:03 Voltage Offset Correction: v_off_ini = 101.953 uV, v_off_fin = 95.243 uV 19:33:04 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -180.256 uVs 19:33:04 Integrated Voltage Measurement... 19:33:12 Voltage Offset Correction: v_off_ini = 107.886 uV, v_off_fin = 122.789 uV 19:33:13 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 192.578 uVs 19:33:13 Integrated Voltage Measurement... 19:33:22 Voltage Offset Correction: v_off_ini = 94.522 uV, v_off_fin = 83.550 uV 19:33:22 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -169.604 uVs 19:33:22 Integrated Voltage Measurement... 19:33:31 Voltage Offset Correction: v_off_ini = 98.286 uV, v_off_fin = 110.887 uV 19:33:31 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 178.856 uVs 19:33:35 I1Y Average and RMS of Measurements VT = -179.754 +- 5.040 uVs I1Y = -35.951 +- 1.008 uTm 19:33:37 Wire Position: X0 = 0.0000 mm, Y0 = -1.0000 mm 19:33:40 Integrated Voltage Measurement... 19:33:49 Voltage Offset Correction: v_off_ini = 80.341 uV, v_off_fin = 82.975 uV 19:33:49 Integrated Voltage Measurement dXYus = -1.0001 mm, dXYds = 0.0000 mm dXY = -1.0001 mm VT = -101.713 uVs 19:33:49 Integrated Voltage Measurement... 19:33:58 Voltage Offset Correction: v_off_ini = 96.753 uV, v_off_fin = 98.413 uV 19:33:58 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 110.395 uVs 19:33:58 Integrated Voltage Measurement... 19:34:07 Voltage Offset Correction: v_off_ini = 92.180 uV, v_off_fin = 83.473 uV 19:34:07 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -106.970 uVs 19:34:07 Integrated Voltage Measurement... 19:34:16 Voltage Offset Correction: v_off_ini = 96.780 uV, v_off_fin = 98.554 uV 19:34:16 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 104.080 uVs 19:34:16 Integrated Voltage Measurement... 19:34:25 Voltage Offset Correction: v_off_ini = 95.216 uV, v_off_fin = 77.685 uV 19:34:25 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -104.554 uVs 19:34:25 Integrated Voltage Measurement... 19:34:34 Voltage Offset Correction: v_off_ini = 86.324 uV, v_off_fin = 97.046 uV 19:34:34 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 97.817 uVs 19:34:37 I2X Average and RMS of Measurements VT = -104.255 +- 2.181 uVs I2X = 9.591 +- 0.201 uTm^2 19:34:39 Wire Position: X0 = 0.0000 mm, Y0 = -1.0000 mm 19:34:42 Integrated Voltage Measurement... 19:34:51 Voltage Offset Correction: v_off_ini = 106.711 uV, v_off_fin = 104.833 uV 19:34:51 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -281.639 uVs 19:34:51 Integrated Voltage Measurement... 19:35:00 Voltage Offset Correction: v_off_ini = 88.743 uV, v_off_fin = 104.197 uV 19:35:00 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 291.188 uVs 19:35:00 Integrated Voltage Measurement... 19:35:09 Voltage Offset Correction: v_off_ini = 103.868 uV, v_off_fin = 102.131 uV 19:35:09 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -293.625 uVs 19:35:09 Integrated Voltage Measurement... 19:35:18 Voltage Offset Correction: v_off_ini = 95.580 uV, v_off_fin = 94.473 uV 19:35:18 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 271.799 uVs 19:35:18 Integrated Voltage Measurement... 19:35:27 Voltage Offset Correction: v_off_ini = 97.354 uV, v_off_fin = 98.431 uV 19:35:27 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -268.901 uVs 19:35:27 Integrated Voltage Measurement... 19:35:36 Voltage Offset Correction: v_off_ini = 79.848 uV, v_off_fin = 83.487 uV 19:35:36 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 272.308 uVs 19:35:38 I2Y Average and RMS of Measurements VT = -279.910 +- 6.751 uVs I2Y = -25.752 +- 0.621 uTm^2 19:35:42 Wire Position: X0 = 0.0000 mm, Y0 = -0.5000 mm 19:35:47 Integrated Voltage Measurement... 19:35:55 Voltage Offset Correction: v_off_ini = 97.198 uV, v_off_fin = 76.054 uV 19:35:56 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -97.546 uVs 19:35:56 Integrated Voltage Measurement... 19:36:05 Voltage Offset Correction: v_off_ini = 82.209 uV, v_off_fin = 91.892 uV 19:36:05 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 103.547 uVs 19:36:05 Integrated Voltage Measurement... 19:36:14 Voltage Offset Correction: v_off_ini = 92.752 uV, v_off_fin = 83.520 uV 19:36:14 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -104.838 uVs 19:36:14 Integrated Voltage Measurement... 19:36:23 Voltage Offset Correction: v_off_ini = 88.460 uV, v_off_fin = 91.844 uV 19:36:23 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.4999 mm dXY = 0.4999 mm VT = 113.519 uVs 19:36:23 Integrated Voltage Measurement... 19:36:32 Voltage Offset Correction: v_off_ini = 78.993 uV, v_off_fin = 79.550 uV 19:36:32 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -109.518 uVs 19:36:32 Integrated Voltage Measurement... 19:36:41 Voltage Offset Correction: v_off_ini = 88.051 uV, v_off_fin = 81.289 uV 19:36:41 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.4999 mm dXY = 0.4999 mm VT = 110.238 uVs 19:36:45 I1X Average and RMS of Measurements VT = -106.534 +- 4.244 uVs I1X = 21.309 +- 0.850 uTm 19:36:47 Wire Position: X0 = 0.0000 mm, Y0 = -0.5000 mm 19:36:52 Integrated Voltage Measurement... 19:37:01 Voltage Offset Correction: v_off_ini = 84.916 uV, v_off_fin = 85.340 uV 19:37:01 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -180.152 uVs 19:37:01 Integrated Voltage Measurement... 19:37:10 Voltage Offset Correction: v_off_ini = 96.333 uV, v_off_fin = 109.187 uV 19:37:10 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 172.204 uVs 19:37:10 Integrated Voltage Measurement... 19:37:19 Voltage Offset Correction: v_off_ini = 98.219 uV, v_off_fin = 86.711 uV 19:37:19 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -185.064 uVs 19:37:19 Integrated Voltage Measurement... 19:37:28 Voltage Offset Correction: v_off_ini = 89.026 uV, v_off_fin = 100.558 uV 19:37:28 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 190.913 uVs 19:37:28 Integrated Voltage Measurement... 19:37:37 Voltage Offset Correction: v_off_ini = 88.205 uV, v_off_fin = 83.147 uV 19:37:37 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -183.207 uVs 19:37:37 Integrated Voltage Measurement... 19:37:46 Voltage Offset Correction: v_off_ini = 87.532 uV, v_off_fin = 102.197 uV 19:37:46 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 177.207 uVs 19:37:50 I1Y Average and RMS of Measurements VT = -181.458 +- 4.902 uVs I1Y = -36.292 +- 0.980 uTm 19:37:52 Wire Position: X0 = 0.0000 mm, Y0 = -0.5000 mm 19:37:55 Integrated Voltage Measurement... 19:38:04 Voltage Offset Correction: v_off_ini = 105.732 uV, v_off_fin = 91.337 uV 19:38:04 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -137.247 uVs 19:38:04 Integrated Voltage Measurement... 19:38:13 Voltage Offset Correction: v_off_ini = 105.319 uV, v_off_fin = 110.218 uV 19:38:13 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 109.369 uVs 19:38:13 Integrated Voltage Measurement... 19:38:22 Voltage Offset Correction: v_off_ini = 109.970 uV, v_off_fin = 103.022 uV 19:38:22 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -130.372 uVs 19:38:22 Integrated Voltage Measurement... 19:38:31 Voltage Offset Correction: v_off_ini = 97.958 uV, v_off_fin = 92.170 uV 19:38:31 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 121.993 uVs 19:38:31 Integrated Voltage Measurement... 19:38:40 Voltage Offset Correction: v_off_ini = 98.163 uV, v_off_fin = 87.645 uV 19:38:40 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -130.875 uVs 19:38:40 Integrated Voltage Measurement... 19:38:49 Voltage Offset Correction: v_off_ini = 104.793 uV, v_off_fin = 100.715 uV 19:38:49 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 131.722 uVs 19:38:52 I2X Average and RMS of Measurements VT = -126.930 +- 3.304 uVs I2X = 11.678 +- 0.304 uTm^2 19:38:54 Wire Position: X0 = 0.0000 mm, Y0 = -0.5000 mm 19:38:57 Integrated Voltage Measurement... 19:39:06 Voltage Offset Correction: v_off_ini = 95.863 uV, v_off_fin = 96.163 uV 19:39:06 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -288.695 uVs 19:39:06 Integrated Voltage Measurement... 19:39:15 Voltage Offset Correction: v_off_ini = 97.789 uV, v_off_fin = 94.058 uV 19:39:15 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 275.835 uVs 19:39:15 Integrated Voltage Measurement... 19:39:24 Voltage Offset Correction: v_off_ini = 87.629 uV, v_off_fin = 83.287 uV 19:39:24 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -274.775 uVs 19:39:24 Integrated Voltage Measurement... 19:39:33 Voltage Offset Correction: v_off_ini = 93.592 uV, v_off_fin = 103.299 uV 19:39:33 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 269.391 uVs 19:39:33 Integrated Voltage Measurement... 19:39:42 Voltage Offset Correction: v_off_ini = 104.825 uV, v_off_fin = 102.784 uV 19:39:42 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -271.084 uVs 19:39:42 Integrated Voltage Measurement... 19:39:51 Voltage Offset Correction: v_off_ini = 94.520 uV, v_off_fin = 98.352 uV 19:39:51 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 278.651 uVs 19:39:53 I2Y Average and RMS of Measurements VT = -276.405 +- 4.296 uVs I2Y = -25.429 +- 0.395 uTm^2 19:39:57 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 19:40:02 Integrated Voltage Measurement... 19:40:11 Voltage Offset Correction: v_off_ini = 102.830 uV, v_off_fin = 89.337 uV 19:40:11 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.4998 mm dXY = -0.4999 mm VT = -94.728 uVs 19:40:11 Integrated Voltage Measurement... 19:40:20 Voltage Offset Correction: v_off_ini = 94.541 uV, v_off_fin = 110.260 uV 19:40:20 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.4999 mm dXY = 0.4999 mm VT = 108.040 uVs 19:40:20 Integrated Voltage Measurement... 19:40:29 Voltage Offset Correction: v_off_ini = 108.851 uV, v_off_fin = 99.388 uV 19:40:29 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -119.396 uVs 19:40:29 Integrated Voltage Measurement... 19:40:38 Voltage Offset Correction: v_off_ini = 102.292 uV, v_off_fin = 94.869 uV 19:40:38 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 100.167 uVs 19:40:38 Integrated Voltage Measurement... 19:40:47 Voltage Offset Correction: v_off_ini = 111.529 uV, v_off_fin = 88.917 uV 19:40:47 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -125.765 uVs 19:40:47 Integrated Voltage Measurement... 19:40:56 Voltage Offset Correction: v_off_ini = 101.157 uV, v_off_fin = 109.118 uV 19:40:56 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.4999 mm dXY = 0.4999 mm VT = 99.739 uVs 19:41:00 I1X Average and RMS of Measurements VT = -107.973 +- 4.814 uVs I1X = 21.598 +- 0.962 uTm 19:41:02 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 19:41:07 Integrated Voltage Measurement... 19:41:16 Voltage Offset Correction: v_off_ini = 107.386 uV, v_off_fin = 102.170 uV 19:41:16 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -170.481 uVs 19:41:16 Integrated Voltage Measurement... 19:41:25 Voltage Offset Correction: v_off_ini = 103.711 uV, v_off_fin = 114.293 uV 19:41:25 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 170.385 uVs 19:41:25 Integrated Voltage Measurement... 19:41:34 Voltage Offset Correction: v_off_ini = 97.732 uV, v_off_fin = 99.997 uV 19:41:34 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -176.792 uVs 19:41:34 Integrated Voltage Measurement... 19:41:43 Voltage Offset Correction: v_off_ini = 107.873 uV, v_off_fin = 120.477 uV 19:41:43 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 180.350 uVs 19:41:43 Integrated Voltage Measurement... 19:41:52 Voltage Offset Correction: v_off_ini = 106.379 uV, v_off_fin = 105.377 uV 19:41:52 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -179.285 uVs 19:41:52 Integrated Voltage Measurement... 19:42:01 Voltage Offset Correction: v_off_ini = 109.520 uV, v_off_fin = 122.193 uV 19:42:01 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 153.123 uVs 19:42:05 I1Y Average and RMS of Measurements VT = -171.736 +- 5.132 uVs I1Y = -34.347 +- 1.026 uTm 19:42:07 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 19:42:10 Integrated Voltage Measurement... 19:42:19 Voltage Offset Correction: v_off_ini = 93.243 uV, v_off_fin = 97.691 uV 19:42:19 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -112.992 uVs 19:42:19 Integrated Voltage Measurement... 19:42:28 Voltage Offset Correction: v_off_ini = 96.329 uV, v_off_fin = 116.121 uV 19:42:28 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 112.806 uVs 19:42:28 Integrated Voltage Measurement... 19:42:37 Voltage Offset Correction: v_off_ini = 118.190 uV, v_off_fin = 104.635 uV 19:42:37 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -127.973 uVs 19:42:37 Integrated Voltage Measurement... 19:42:46 Voltage Offset Correction: v_off_ini = 113.135 uV, v_off_fin = 125.913 uV 19:42:46 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 122.352 uVs 19:42:46 Integrated Voltage Measurement... 19:42:55 Voltage Offset Correction: v_off_ini = 122.188 uV, v_off_fin = 106.332 uV 19:42:55 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -118.591 uVs 19:42:55 Integrated Voltage Measurement... 19:43:04 Voltage Offset Correction: v_off_ini = 124.252 uV, v_off_fin = 129.750 uV 19:43:04 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 137.001 uVs 19:43:07 I2X Average and RMS of Measurements VT = -121.953 +- 6.491 uVs I2X = 11.220 +- 0.597 uTm^2 19:43:09 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 19:43:12 Integrated Voltage Measurement... 19:43:21 Voltage Offset Correction: v_off_ini = 139.797 uV, v_off_fin = 117.209 uV 19:43:21 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -289.865 uVs 19:43:21 Integrated Voltage Measurement... 19:43:30 Voltage Offset Correction: v_off_ini = 111.546 uV, v_off_fin = 112.505 uV 19:43:30 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 276.954 uVs 19:43:30 Integrated Voltage Measurement... 19:43:39 Voltage Offset Correction: v_off_ini = 100.875 uV, v_off_fin = 107.371 uV 19:43:39 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -265.328 uVs 19:43:39 Integrated Voltage Measurement... 19:43:48 Voltage Offset Correction: v_off_ini = 105.384 uV, v_off_fin = 110.862 uV 19:43:48 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 282.800 uVs 19:43:48 Integrated Voltage Measurement... 19:43:57 Voltage Offset Correction: v_off_ini = 116.495 uV, v_off_fin = 105.753 uV 19:43:57 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -279.272 uVs 19:43:57 Integrated Voltage Measurement... 19:44:06 Voltage Offset Correction: v_off_ini = 103.999 uV, v_off_fin = 112.784 uV 19:44:06 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 275.952 uVs 19:44:08 I2Y Average and RMS of Measurements VT = -278.362 +- 3.852 uVs I2Y = -25.609 +- 0.354 uTm^2 19:44:13 Wire Position: X0 = 0.0000 mm, Y0 = 0.5000 mm 19:44:17 Integrated Voltage Measurement... 19:44:26 Voltage Offset Correction: v_off_ini = 91.823 uV, v_off_fin = 94.297 uV 19:44:26 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -105.415 uVs 19:44:26 Integrated Voltage Measurement... 19:44:35 Voltage Offset Correction: v_off_ini = 98.731 uV, v_off_fin = 107.937 uV 19:44:35 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 127.975 uVs 19:44:35 Integrated Voltage Measurement... 19:44:44 Voltage Offset Correction: v_off_ini = 108.909 uV, v_off_fin = 103.285 uV 19:44:44 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -116.736 uVs 19:44:44 Integrated Voltage Measurement... 19:44:53 Voltage Offset Correction: v_off_ini = 115.008 uV, v_off_fin = 110.041 uV 19:44:53 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 115.412 uVs 19:44:53 Integrated Voltage Measurement... 19:45:02 Voltage Offset Correction: v_off_ini = 115.921 uV, v_off_fin = 112.540 uV 19:45:02 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -112.074 uVs 19:45:02 Integrated Voltage Measurement... 19:45:11 Voltage Offset Correction: v_off_ini = 118.108 uV, v_off_fin = 115.260 uV 19:45:11 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 118.403 uVs 19:45:15 I1X Average and RMS of Measurements VT = -116.003 +- 0.597 uVs I1X = 23.201 +- 0.119 uTm 19:45:18 Wire Position: X0 = 0.0000 mm, Y0 = 0.5000 mm 19:45:22 Integrated Voltage Measurement... 19:45:31 Voltage Offset Correction: v_off_ini = 120.472 uV, v_off_fin = 121.161 uV 19:45:31 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -155.634 uVs 19:45:31 Integrated Voltage Measurement... 19:45:40 Voltage Offset Correction: v_off_ini = 123.723 uV, v_off_fin = 133.226 uV 19:45:40 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 189.574 uVs 19:45:40 Integrated Voltage Measurement... 19:45:49 Voltage Offset Correction: v_off_ini = 131.455 uV, v_off_fin = 126.406 uV 19:45:49 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -170.007 uVs 19:45:49 Integrated Voltage Measurement... 19:45:58 Voltage Offset Correction: v_off_ini = 129.450 uV, v_off_fin = 142.284 uV 19:45:58 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 194.307 uVs 19:45:58 Integrated Voltage Measurement... 19:46:07 Voltage Offset Correction: v_off_ini = 123.349 uV, v_off_fin = 128.027 uV 19:46:07 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -163.875 uVs 19:46:07 Integrated Voltage Measurement... 19:46:16 Voltage Offset Correction: v_off_ini = 140.113 uV, v_off_fin = 130.893 uV 19:46:16 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 174.463 uVs 19:46:20 I1Y Average and RMS of Measurements VT = -174.643 +- 5.495 uVs I1Y = -34.926 +- 1.101 uTm 19:46:23 Wire Position: X0 = 0.0000 mm, Y0 = 0.5000 mm 19:46:25 Integrated Voltage Measurement... 19:46:34 Voltage Offset Correction: v_off_ini = 105.997 uV, v_off_fin = 104.531 uV 19:46:34 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -127.963 uVs 19:46:34 Integrated Voltage Measurement... 19:46:43 Voltage Offset Correction: v_off_ini = 114.654 uV, v_off_fin = 118.755 uV 19:46:43 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 125.007 uVs 19:46:43 Integrated Voltage Measurement... 19:46:52 Voltage Offset Correction: v_off_ini = 110.520 uV, v_off_fin = 106.481 uV 19:46:52 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -102.459 uVs 19:46:52 Integrated Voltage Measurement... 19:47:01 Voltage Offset Correction: v_off_ini = 128.491 uV, v_off_fin = 119.300 uV 19:47:01 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 113.071 uVs 19:47:01 Integrated Voltage Measurement... 19:47:10 Voltage Offset Correction: v_off_ini = 104.955 uV, v_off_fin = 105.891 uV 19:47:10 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -122.433 uVs 19:47:11 Integrated Voltage Measurement... 19:47:19 Voltage Offset Correction: v_off_ini = 117.414 uV, v_off_fin = 120.700 uV 19:47:19 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 120.608 uVs 19:47:22 I2X Average and RMS of Measurements VT = -118.590 +- 7.918 uVs I2X = 10.910 +- 0.728 uTm^2 19:47:24 Wire Position: X0 = 0.0000 mm, Y0 = 0.5000 mm 19:47:27 Integrated Voltage Measurement... 19:47:36 Voltage Offset Correction: v_off_ini = 126.567 uV, v_off_fin = 111.554 uV 19:47:36 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = -254.327 uVs 19:47:36 Integrated Voltage Measurement... 19:47:45 Voltage Offset Correction: v_off_ini = 126.903 uV, v_off_fin = 115.961 uV 19:47:45 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 261.531 uVs 19:47:45 Integrated Voltage Measurement... 19:47:54 Voltage Offset Correction: v_off_ini = 102.454 uV, v_off_fin = 109.692 uV 19:47:54 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -262.749 uVs 19:47:54 Integrated Voltage Measurement... 19:48:03 Voltage Offset Correction: v_off_ini = 121.779 uV, v_off_fin = 119.365 uV 19:48:03 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 269.877 uVs 19:48:03 Integrated Voltage Measurement... 19:48:12 Voltage Offset Correction: v_off_ini = 125.264 uV, v_off_fin = 119.294 uV 19:48:12 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -275.210 uVs 19:48:12 Integrated Voltage Measurement... 19:48:21 Voltage Offset Correction: v_off_ini = 116.182 uV, v_off_fin = 129.433 uV 19:48:21 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 262.923 uVs 19:48:24 I2Y Average and RMS of Measurements VT = -264.436 +- 4.737 uVs I2Y = -24.329 +- 0.435 uTm^2 19:48:28 Wire Position: X0 = 0.0000 mm, Y0 = 1.0000 mm 19:48:32 Integrated Voltage Measurement... 19:48:41 Voltage Offset Correction: v_off_ini = 119.636 uV, v_off_fin = 116.658 uV 19:48:41 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -115.145 uVs 19:48:41 Integrated Voltage Measurement... 19:48:50 Voltage Offset Correction: v_off_ini = 118.729 uV, v_off_fin = 114.278 uV 19:48:50 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 91.478 uVs 19:48:50 Integrated Voltage Measurement... 19:48:59 Voltage Offset Correction: v_off_ini = 127.703 uV, v_off_fin = 129.381 uV 19:48:59 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -85.595 uVs 19:48:59 Integrated Voltage Measurement... 19:49:08 Voltage Offset Correction: v_off_ini = 123.284 uV, v_off_fin = 137.165 uV 19:49:08 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 116.172 uVs 19:49:08 Integrated Voltage Measurement... 19:49:17 Voltage Offset Correction: v_off_ini = 130.921 uV, v_off_fin = 115.451 uV 19:49:17 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -92.014 uVs 19:49:17 Integrated Voltage Measurement... 19:49:26 Voltage Offset Correction: v_off_ini = 118.743 uV, v_off_fin = 121.450 uV 19:49:26 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 111.385 uVs 19:49:30 I1X Average and RMS of Measurements VT = -101.965 +- 1.009 uVs I1X = 20.394 +- 0.201 uTm 19:49:33 Wire Position: X0 = 0.0000 mm, Y0 = 1.0000 mm 19:49:37 Integrated Voltage Measurement... 19:49:46 Voltage Offset Correction: v_off_ini = 115.177 uV, v_off_fin = 107.611 uV 19:49:46 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -164.290 uVs 19:49:46 Integrated Voltage Measurement... 19:49:55 Voltage Offset Correction: v_off_ini = 120.918 uV, v_off_fin = 136.350 uV 19:49:55 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 176.667 uVs 19:49:55 Integrated Voltage Measurement... 19:50:04 Voltage Offset Correction: v_off_ini = 115.860 uV, v_off_fin = 111.203 uV 19:50:04 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -177.047 uVs 19:50:04 Integrated Voltage Measurement... 19:50:13 Voltage Offset Correction: v_off_ini = 126.956 uV, v_off_fin = 140.130 uV 19:50:13 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 170.406 uVs 19:50:13 Integrated Voltage Measurement... 19:50:22 Voltage Offset Correction: v_off_ini = 129.974 uV, v_off_fin = 105.405 uV 19:50:22 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -172.912 uVs 19:50:22 Integrated Voltage Measurement... 19:50:31 Voltage Offset Correction: v_off_ini = 133.131 uV, v_off_fin = 141.367 uV 19:50:31 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 176.646 uVs 19:50:35 I1Y Average and RMS of Measurements VT = -172.995 +- 1.830 uVs I1Y = -34.597 +- 0.367 uTm 19:50:38 Wire Position: X0 = 0.0000 mm, Y0 = 1.0000 mm 19:50:40 Integrated Voltage Measurement... 19:50:49 Voltage Offset Correction: v_off_ini = 128.801 uV, v_off_fin = 124.275 uV 19:50:49 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -136.401 uVs 19:50:50 Integrated Voltage Measurement... 19:50:58 Voltage Offset Correction: v_off_ini = 121.917 uV, v_off_fin = 125.796 uV 19:50:59 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 118.838 uVs 19:50:59 Integrated Voltage Measurement... 19:51:08 Voltage Offset Correction: v_off_ini = 105.998 uV, v_off_fin = 90.318 uV 19:51:08 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -106.282 uVs 19:51:08 Integrated Voltage Measurement... 19:51:17 Voltage Offset Correction: v_off_ini = 107.810 uV, v_off_fin = 123.101 uV 19:51:17 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 109.855 uVs 19:51:17 Integrated Voltage Measurement... 19:51:26 Voltage Offset Correction: v_off_ini = 107.174 uV, v_off_fin = 101.787 uV 19:51:26 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -109.603 uVs 19:51:26 Integrated Voltage Measurement... 19:51:35 Voltage Offset Correction: v_off_ini = 106.999 uV, v_off_fin = 115.498 uV 19:51:35 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 108.797 uVs 19:51:37 I2X Average and RMS of Measurements VT = -114.963 +- 8.962 uVs I2X = 10.577 +- 0.824 uTm^2 19:51:40 Wire Position: X0 = 0.0000 mm, Y0 = 1.0000 mm 19:51:42 Integrated Voltage Measurement... 19:51:51 Voltage Offset Correction: v_off_ini = 110.092 uV, v_off_fin = 104.144 uV 19:51:51 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -272.526 uVs 19:51:51 Integrated Voltage Measurement... 19:52:00 Voltage Offset Correction: v_off_ini = 106.100 uV, v_off_fin = 111.909 uV 19:52:00 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 271.776 uVs 19:52:00 Integrated Voltage Measurement... 19:52:09 Voltage Offset Correction: v_off_ini = 113.778 uV, v_off_fin = 105.940 uV 19:52:09 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -271.329 uVs 19:52:09 Integrated Voltage Measurement... 19:52:18 Voltage Offset Correction: v_off_ini = 101.674 uV, v_off_fin = 115.431 uV 19:52:18 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 276.650 uVs 19:52:18 Integrated Voltage Measurement... 19:52:27 Voltage Offset Correction: v_off_ini = 106.153 uV, v_off_fin = 113.616 uV 19:52:27 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -262.753 uVs 19:52:27 Integrated Voltage Measurement... 19:52:36 Voltage Offset Correction: v_off_ini = 113.369 uV, v_off_fin = 110.367 uV 19:52:36 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 285.385 uVs 19:52:39 I2Y Average and RMS of Measurements VT = -273.403 +- 0.886 uVs I2Y = -25.153 +- 0.082 uTm^2 19:52:56 Wire Position: X0 = -2.0000 mm, Y0 = 0.0000 mm 19:53:00 Integrated Voltage Measurement... 19:53:09 Voltage Offset Correction: v_off_ini = 113.434 uV, v_off_fin = 97.987 uV 19:53:09 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -130.903 uVs 19:53:09 Integrated Voltage Measurement... 19:53:18 Voltage Offset Correction: v_off_ini = 126.253 uV, v_off_fin = 124.488 uV 19:53:18 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 112.932 uVs 19:53:18 Integrated Voltage Measurement... 19:53:27 Voltage Offset Correction: v_off_ini = 122.896 uV, v_off_fin = 123.149 uV 19:53:27 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -122.218 uVs 19:53:27 Integrated Voltage Measurement... 19:53:36 Voltage Offset Correction: v_off_ini = 111.534 uV, v_off_fin = 113.764 uV 19:53:36 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 126.022 uVs 19:53:36 Integrated Voltage Measurement... 19:53:45 Voltage Offset Correction: v_off_ini = 116.476 uV, v_off_fin = 109.190 uV 19:53:45 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -132.207 uVs 19:53:45 Integrated Voltage Measurement... 19:53:54 Voltage Offset Correction: v_off_ini = 111.990 uV, v_off_fin = 119.638 uV 19:53:54 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 130.097 uVs 19:53:58 I1X Average and RMS of Measurements VT = -125.730 +- 3.938 uVs I1X = 25.147 +- 0.788 uTm 19:54:01 Wire Position: X0 = -2.0000 mm, Y0 = 0.0000 mm 19:54:05 Integrated Voltage Measurement... 19:54:14 Voltage Offset Correction: v_off_ini = 107.288 uV, v_off_fin = 101.494 uV 19:54:14 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -170.375 uVs 19:54:14 Integrated Voltage Measurement... 19:54:23 Voltage Offset Correction: v_off_ini = 107.480 uV, v_off_fin = 128.365 uV 19:54:23 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 177.372 uVs 19:54:23 Integrated Voltage Measurement... 19:54:32 Voltage Offset Correction: v_off_ini = 122.670 uV, v_off_fin = 106.102 uV 19:54:32 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -182.883 uVs 19:54:32 Integrated Voltage Measurement... 19:54:41 Voltage Offset Correction: v_off_ini = 110.535 uV, v_off_fin = 122.895 uV 19:54:41 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 186.485 uVs 19:54:41 Integrated Voltage Measurement... 19:54:50 Voltage Offset Correction: v_off_ini = 117.496 uV, v_off_fin = 108.057 uV 19:54:50 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -199.062 uVs 19:54:50 Integrated Voltage Measurement... 19:54:59 Voltage Offset Correction: v_off_ini = 106.340 uV, v_off_fin = 117.533 uV 19:54:59 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 165.151 uVs 19:55:03 I1Y Average and RMS of Measurements VT = -180.221 +- 4.610 uVs I1Y = -36.044 +- 0.922 uTm 19:55:09 Wire Position: X0 = -2.0000 mm, Y0 = 0.0000 mm 19:55:11 Integrated Voltage Measurement... 19:55:20 Voltage Offset Correction: v_off_ini = 88.549 uV, v_off_fin = 91.189 uV 19:55:20 Integrated Voltage Measurement dXYus = -1.0001 mm, dXYds = 0.0000 mm dXY = -1.0001 mm VT = -124.521 uVs 19:55:20 Integrated Voltage Measurement... 19:55:29 Voltage Offset Correction: v_off_ini = 88.397 uV, v_off_fin = 109.589 uV 19:55:29 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 143.223 uVs 19:55:29 Integrated Voltage Measurement... 19:55:38 Voltage Offset Correction: v_off_ini = 119.453 uV, v_off_fin = 102.062 uV 19:55:38 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -148.007 uVs 19:55:38 Integrated Voltage Measurement... 19:55:47 Voltage Offset Correction: v_off_ini = 102.115 uV, v_off_fin = 114.954 uV 19:55:47 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 156.169 uVs 19:55:47 Integrated Voltage Measurement... 19:55:56 Voltage Offset Correction: v_off_ini = 118.574 uV, v_off_fin = 109.716 uV 19:55:56 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -126.991 uVs 19:55:56 Integrated Voltage Measurement... 19:56:05 Voltage Offset Correction: v_off_ini = 112.932 uV, v_off_fin = 126.022 uV 19:56:05 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 111.136 uVs 19:56:08 I2X Average and RMS of Measurements VT = -135.008 +- 13.506 uVs I2X = 12.421 +- 1.243 uTm^2 19:56:13 Wire Position: X0 = -2.0000 mm, Y0 = 0.0000 mm 19:56:16 Integrated Voltage Measurement... 19:56:25 Voltage Offset Correction: v_off_ini = 109.828 uV, v_off_fin = 120.676 uV 19:56:25 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -259.757 uVs 19:56:25 Integrated Voltage Measurement... 19:56:34 Voltage Offset Correction: v_off_ini = 108.521 uV, v_off_fin = 115.247 uV 19:56:34 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 277.897 uVs 19:56:34 Integrated Voltage Measurement... 19:56:43 Voltage Offset Correction: v_off_ini = 117.176 uV, v_off_fin = 125.924 uV 19:56:43 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -270.771 uVs 19:56:43 Integrated Voltage Measurement... 19:56:52 Voltage Offset Correction: v_off_ini = 123.548 uV, v_off_fin = 122.952 uV 19:56:52 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 281.189 uVs 19:56:52 Integrated Voltage Measurement... 19:57:01 Voltage Offset Correction: v_off_ini = 125.643 uV, v_off_fin = 134.405 uV 19:57:01 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -281.041 uVs 19:57:01 Integrated Voltage Measurement... 19:57:10 Voltage Offset Correction: v_off_ini = 130.127 uV, v_off_fin = 132.184 uV 19:57:10 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 258.776 uVs 19:57:12 I2Y Average and RMS of Measurements VT = -271.572 +- 3.148 uVs I2Y = -24.985 +- 0.290 uTm^2 19:57:17 Wire Position: X0 = -1.0000 mm, Y0 = 0.0000 mm 19:57:21 Integrated Voltage Measurement... 19:57:30 Voltage Offset Correction: v_off_ini = 123.240 uV, v_off_fin = 117.071 uV 19:57:30 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.4998 mm dXY = -0.4999 mm VT = -113.995 uVs 19:57:30 Integrated Voltage Measurement... 19:57:39 Voltage Offset Correction: v_off_ini = 123.556 uV, v_off_fin = 133.646 uV 19:57:39 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 98.064 uVs 19:57:39 Integrated Voltage Measurement... 19:57:48 Voltage Offset Correction: v_off_ini = 119.005 uV, v_off_fin = 135.731 uV 19:57:48 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -118.774 uVs 19:57:48 Integrated Voltage Measurement... 19:57:57 Voltage Offset Correction: v_off_ini = 140.792 uV, v_off_fin = 134.596 uV 19:57:57 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 113.462 uVs 19:57:57 Integrated Voltage Measurement... 19:58:06 Voltage Offset Correction: v_off_ini = 123.664 uV, v_off_fin = 111.365 uV 19:58:06 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -108.610 uVs 19:58:06 Integrated Voltage Measurement... 19:58:15 Voltage Offset Correction: v_off_ini = 124.626 uV, v_off_fin = 128.385 uV 19:58:15 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 129.398 uVs 19:58:19 I1X Average and RMS of Measurements VT = -113.717 +- 5.562 uVs I1X = 22.745 +- 1.111 uTm 19:58:22 Wire Position: X0 = -1.0000 mm, Y0 = 0.0000 mm 19:58:26 Integrated Voltage Measurement... 19:58:35 Voltage Offset Correction: v_off_ini = 124.171 uV, v_off_fin = 110.697 uV 19:58:35 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -207.504 uVs 19:58:35 Integrated Voltage Measurement... 19:58:44 Voltage Offset Correction: v_off_ini = 126.306 uV, v_off_fin = 132.281 uV 19:58:44 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 186.405 uVs 19:58:44 Integrated Voltage Measurement... 19:58:53 Voltage Offset Correction: v_off_ini = 122.434 uV, v_off_fin = 114.301 uV 19:58:53 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -191.646 uVs 19:58:53 Integrated Voltage Measurement... 19:59:02 Voltage Offset Correction: v_off_ini = 117.829 uV, v_off_fin = 127.465 uV 19:59:02 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 190.256 uVs 19:59:02 Integrated Voltage Measurement... 19:59:11 Voltage Offset Correction: v_off_ini = 125.605 uV, v_off_fin = 110.282 uV 19:59:11 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -179.604 uVs 19:59:11 Integrated Voltage Measurement... 19:59:20 Voltage Offset Correction: v_off_ini = 123.505 uV, v_off_fin = 127.595 uV 19:59:20 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 183.810 uVs 19:59:24 I1Y Average and RMS of Measurements VT = -189.871 +- 6.271 uVs I1Y = -37.974 +- 1.254 uTm 19:59:30 Wire Position: X0 = -1.0000 mm, Y0 = 0.0000 mm 19:59:32 Integrated Voltage Measurement... 19:59:41 Voltage Offset Correction: v_off_ini = 118.266 uV, v_off_fin = 127.102 uV 19:59:41 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -126.567 uVs 19:59:41 Integrated Voltage Measurement... 19:59:50 Voltage Offset Correction: v_off_ini = 124.813 uV, v_off_fin = 138.035 uV 19:59:50 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 124.466 uVs 19:59:50 Integrated Voltage Measurement... 19:59:59 Voltage Offset Correction: v_off_ini = 132.614 uV, v_off_fin = 124.601 uV 19:59:59 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -116.098 uVs 19:59:59 Integrated Voltage Measurement... 20:00:08 Voltage Offset Correction: v_off_ini = 129.161 uV, v_off_fin = 152.236 uV 20:00:08 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 139.222 uVs 20:00:08 Integrated Voltage Measurement... 20:00:17 Voltage Offset Correction: v_off_ini = 144.271 uV, v_off_fin = 135.110 uV 20:00:17 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -132.435 uVs 20:00:17 Integrated Voltage Measurement... 20:00:26 Voltage Offset Correction: v_off_ini = 142.867 uV, v_off_fin = 142.994 uV 20:00:26 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 144.366 uVs 20:00:29 I2X Average and RMS of Measurements VT = -130.526 +- 5.637 uVs I2X = 12.009 +- 0.518 uTm^2 20:00:34 Wire Position: X0 = -1.0000 mm, Y0 = 0.0000 mm 20:00:37 Integrated Voltage Measurement... 20:00:46 Voltage Offset Correction: v_off_ini = 133.511 uV, v_off_fin = 123.974 uV 20:00:46 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -249.613 uVs 20:00:46 Integrated Voltage Measurement... 20:00:55 Voltage Offset Correction: v_off_ini = 132.148 uV, v_off_fin = 131.398 uV 20:00:55 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 267.957 uVs 20:00:55 Integrated Voltage Measurement... 20:01:04 Voltage Offset Correction: v_off_ini = 146.248 uV, v_off_fin = 120.111 uV 20:01:04 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -264.124 uVs 20:01:04 Integrated Voltage Measurement... 20:01:13 Voltage Offset Correction: v_off_ini = 141.584 uV, v_off_fin = 134.829 uV 20:01:13 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 250.694 uVs 20:01:13 Integrated Voltage Measurement... 20:01:22 Voltage Offset Correction: v_off_ini = 135.156 uV, v_off_fin = 128.626 uV 20:01:22 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -267.046 uVs 20:01:22 Integrated Voltage Measurement... 20:01:31 Voltage Offset Correction: v_off_ini = 128.504 uV, v_off_fin = 137.199 uV 20:01:31 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 270.955 uVs 20:01:33 I2Y Average and RMS of Measurements VT = -261.732 +- 5.171 uVs I2Y = -24.079 +- 0.476 uTm^2 20:01:38 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 20:01:42 Integrated Voltage Measurement... 20:01:51 Voltage Offset Correction: v_off_ini = 107.885 uV, v_off_fin = 116.263 uV 20:01:51 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -97.543 uVs 20:01:51 Integrated Voltage Measurement... 20:02:00 Voltage Offset Correction: v_off_ini = 133.795 uV, v_off_fin = 128.005 uV 20:02:00 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 121.044 uVs 20:02:00 Integrated Voltage Measurement... 20:02:09 Voltage Offset Correction: v_off_ini = 127.751 uV, v_off_fin = 124.047 uV 20:02:09 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -116.621 uVs 20:02:09 Integrated Voltage Measurement... 20:02:18 Voltage Offset Correction: v_off_ini = 132.220 uV, v_off_fin = 133.336 uV 20:02:18 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.4999 mm dXY = 0.4999 mm VT = 115.885 uVs 20:02:18 Integrated Voltage Measurement... 20:02:27 Voltage Offset Correction: v_off_ini = 131.079 uV, v_off_fin = 132.778 uV 20:02:27 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -131.069 uVs 20:02:27 Integrated Voltage Measurement... 20:02:36 Voltage Offset Correction: v_off_ini = 139.249 uV, v_off_fin = 135.177 uV 20:02:36 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 137.825 uVs 20:02:40 I1X Average and RMS of Measurements VT = -119.998 +- 10.605 uVs I1X = 24.000 +- 2.121 uTm 20:02:43 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 20:02:47 Integrated Voltage Measurement... 20:02:56 Voltage Offset Correction: v_off_ini = 125.921 uV, v_off_fin = 117.012 uV 20:02:56 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = -162.812 uVs 20:02:56 Integrated Voltage Measurement... 20:03:05 Voltage Offset Correction: v_off_ini = 135.097 uV, v_off_fin = 143.718 uV 20:03:05 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 189.415 uVs 20:03:05 Integrated Voltage Measurement... 20:03:14 Voltage Offset Correction: v_off_ini = 127.477 uV, v_off_fin = 132.662 uV 20:03:14 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -176.432 uVs 20:03:14 Integrated Voltage Measurement... 20:03:23 Voltage Offset Correction: v_off_ini = 123.158 uV, v_off_fin = 140.467 uV 20:03:23 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 168.198 uVs 20:03:23 Integrated Voltage Measurement... 20:03:32 Voltage Offset Correction: v_off_ini = 131.006 uV, v_off_fin = 124.741 uV 20:03:32 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -173.635 uVs 20:03:32 Integrated Voltage Measurement... 20:03:41 Voltage Offset Correction: v_off_ini = 130.224 uV, v_off_fin = 135.089 uV 20:03:41 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 175.671 uVs 20:03:45 I1Y Average and RMS of Measurements VT = -174.360 +- 1.565 uVs I1Y = -34.873 +- 0.314 uTm 20:03:50 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 20:03:53 Integrated Voltage Measurement... 20:04:02 Voltage Offset Correction: v_off_ini = 118.356 uV, v_off_fin = 119.631 uV 20:04:02 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -123.871 uVs 20:04:02 Integrated Voltage Measurement... 20:04:11 Voltage Offset Correction: v_off_ini = 114.054 uV, v_off_fin = 122.737 uV 20:04:11 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 148.591 uVs 20:04:11 Integrated Voltage Measurement... 20:04:20 Voltage Offset Correction: v_off_ini = 119.743 uV, v_off_fin = 121.778 uV 20:04:20 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -130.424 uVs 20:04:20 Integrated Voltage Measurement... 20:04:29 Voltage Offset Correction: v_off_ini = 133.374 uV, v_off_fin = 138.863 uV 20:04:29 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 128.580 uVs 20:04:29 Integrated Voltage Measurement... 20:04:38 Voltage Offset Correction: v_off_ini = 141.854 uV, v_off_fin = 127.234 uV 20:04:38 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -139.462 uVs 20:04:38 Integrated Voltage Measurement... 20:04:47 Voltage Offset Correction: v_off_ini = 138.854 uV, v_off_fin = 134.543 uV 20:04:47 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 131.773 uVs 20:04:50 I2X Average and RMS of Measurements VT = -133.783 +- 3.038 uVs I2X = 12.308 +- 0.279 uTm^2 20:04:55 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 20:04:57 Integrated Voltage Measurement... 20:05:06 Voltage Offset Correction: v_off_ini = 134.145 uV, v_off_fin = 134.609 uV 20:05:06 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -263.947 uVs 20:05:06 Integrated Voltage Measurement... 20:05:15 Voltage Offset Correction: v_off_ini = 127.778 uV, v_off_fin = 121.970 uV 20:05:15 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 264.571 uVs 20:05:15 Integrated Voltage Measurement... 20:05:24 Voltage Offset Correction: v_off_ini = 117.097 uV, v_off_fin = 118.173 uV 20:05:24 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -262.677 uVs 20:05:24 Integrated Voltage Measurement... 20:05:33 Voltage Offset Correction: v_off_ini = 120.027 uV, v_off_fin = 122.646 uV 20:05:33 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 253.107 uVs 20:05:33 Integrated Voltage Measurement... 20:05:42 Voltage Offset Correction: v_off_ini = 115.515 uV, v_off_fin = 122.738 uV 20:05:42 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -259.712 uVs 20:05:42 Integrated Voltage Measurement... 20:05:51 Voltage Offset Correction: v_off_ini = 118.516 uV, v_off_fin = 125.161 uV 20:05:51 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 267.548 uVs 20:05:54 I2Y Average and RMS of Measurements VT = -261.927 +- 2.865 uVs I2Y = -24.097 +- 0.264 uTm^2 20:05:58 Wire Position: X0 = 1.0000 mm, Y0 = 0.0000 mm 20:06:03 Integrated Voltage Measurement... 20:06:12 Voltage Offset Correction: v_off_ini = 109.763 uV, v_off_fin = 110.449 uV 20:06:12 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -127.258 uVs 20:06:12 Integrated Voltage Measurement... 20:06:21 Voltage Offset Correction: v_off_ini = 115.151 uV, v_off_fin = 124.007 uV 20:06:21 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 108.715 uVs 20:06:21 Integrated Voltage Measurement... 20:06:30 Voltage Offset Correction: v_off_ini = 95.905 uV, v_off_fin = 106.222 uV 20:06:30 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -111.058 uVs 20:06:30 Integrated Voltage Measurement... 20:06:39 Voltage Offset Correction: v_off_ini = 122.021 uV, v_off_fin = 122.471 uV 20:06:39 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 120.591 uVs 20:06:39 Integrated Voltage Measurement... 20:06:48 Voltage Offset Correction: v_off_ini = 116.751 uV, v_off_fin = 103.031 uV 20:06:48 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -121.668 uVs 20:06:48 Integrated Voltage Measurement... 20:06:57 Voltage Offset Correction: v_off_ini = 110.846 uV, v_off_fin = 125.207 uV 20:06:57 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 127.909 uVs 20:07:01 I1X Average and RMS of Measurements VT = -119.533 +- 3.820 uVs I1X = 23.907 +- 0.764 uTm 20:07:03 Wire Position: X0 = 1.0000 mm, Y0 = 0.0000 mm 20:07:08 Integrated Voltage Measurement... 20:07:17 Voltage Offset Correction: v_off_ini = 109.944 uV, v_off_fin = 97.107 uV 20:07:17 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -176.121 uVs 20:07:17 Integrated Voltage Measurement... 20:07:26 Voltage Offset Correction: v_off_ini = 109.681 uV, v_off_fin = 119.039 uV 20:07:26 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 197.054 uVs 20:07:26 Integrated Voltage Measurement... 20:07:35 Voltage Offset Correction: v_off_ini = 119.827 uV, v_off_fin = 110.086 uV 20:07:35 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -186.761 uVs 20:07:35 Integrated Voltage Measurement... 20:07:44 Voltage Offset Correction: v_off_ini = 114.718 uV, v_off_fin = 139.075 uV 20:07:44 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 189.040 uVs 20:07:44 Integrated Voltage Measurement... 20:07:53 Voltage Offset Correction: v_off_ini = 123.425 uV, v_off_fin = 105.928 uV 20:07:53 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -186.575 uVs 20:07:53 Integrated Voltage Measurement... 20:08:02 Voltage Offset Correction: v_off_ini = 110.604 uV, v_off_fin = 118.394 uV 20:08:02 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 189.586 uVs 20:08:06 I1Y Average and RMS of Measurements VT = -187.523 +- 0.666 uVs I1Y = -37.505 +- 0.133 uTm 20:08:11 Wire Position: X0 = 1.0000 mm, Y0 = 0.0000 mm 20:08:14 Integrated Voltage Measurement... 20:08:23 Voltage Offset Correction: v_off_ini = 114.651 uV, v_off_fin = 97.394 uV 20:08:23 Integrated Voltage Measurement dXYus = -1.0001 mm, dXYds = 0.0000 mm dXY = -1.0001 mm VT = -90.207 uVs 20:08:23 Integrated Voltage Measurement... 20:08:32 Voltage Offset Correction: v_off_ini = 124.237 uV, v_off_fin = 137.856 uV 20:08:32 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 152.747 uVs 20:08:32 Integrated Voltage Measurement... 20:08:41 Voltage Offset Correction: v_off_ini = 126.130 uV, v_off_fin = 114.797 uV 20:08:41 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -134.514 uVs 20:08:41 Integrated Voltage Measurement... 20:08:50 Voltage Offset Correction: v_off_ini = 129.729 uV, v_off_fin = 142.874 uV 20:08:50 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 149.984 uVs 20:08:50 Integrated Voltage Measurement... 20:08:59 Voltage Offset Correction: v_off_ini = 135.260 uV, v_off_fin = 125.103 uV 20:08:59 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -133.888 uVs 20:08:59 Integrated Voltage Measurement... 20:09:08 Voltage Offset Correction: v_off_ini = 146.684 uV, v_off_fin = 148.406 uV 20:09:08 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 118.576 uVs 20:09:10 I2X Average and RMS of Measurements VT = -129.986 +- 8.886 uVs I2X = 11.959 +- 0.818 uTm^2 20:09:15 Wire Position: X0 = 1.0000 mm, Y0 = 0.0000 mm 20:09:18 Integrated Voltage Measurement... 20:09:27 Voltage Offset Correction: v_off_ini = 131.751 uV, v_off_fin = 136.665 uV 20:09:27 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -271.369 uVs 20:09:27 Integrated Voltage Measurement... 20:09:36 Voltage Offset Correction: v_off_ini = 141.787 uV, v_off_fin = 136.642 uV 20:09:36 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 264.581 uVs 20:09:36 Integrated Voltage Measurement... 20:09:45 Voltage Offset Correction: v_off_ini = 142.608 uV, v_off_fin = 142.868 uV 20:09:45 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -263.506 uVs 20:09:45 Integrated Voltage Measurement... 20:09:54 Voltage Offset Correction: v_off_ini = 145.252 uV, v_off_fin = 142.858 uV 20:09:54 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 275.328 uVs 20:09:54 Integrated Voltage Measurement... 20:10:03 Voltage Offset Correction: v_off_ini = 125.950 uV, v_off_fin = 140.883 uV 20:10:03 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -249.898 uVs 20:10:03 Integrated Voltage Measurement... 20:10:12 Voltage Offset Correction: v_off_ini = 137.184 uV, v_off_fin = 128.327 uV 20:10:12 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 262.113 uVs 20:10:14 I2Y Average and RMS of Measurements VT = -264.466 +- 6.011 uVs I2Y = -24.331 +- 0.553 uTm^2 20:10:19 Wire Position: X0 = 2.0000 mm, Y0 = 0.0000 mm 20:10:23 Integrated Voltage Measurement... 20:10:32 Voltage Offset Correction: v_off_ini = 142.279 uV, v_off_fin = 139.665 uV 20:10:32 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.4998 mm dXY = -0.4999 mm VT = -113.507 uVs 20:10:32 Integrated Voltage Measurement... 20:10:41 Voltage Offset Correction: v_off_ini = 140.428 uV, v_off_fin = 146.719 uV 20:10:41 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = 108.390 uVs 20:10:41 Integrated Voltage Measurement... 20:10:50 Voltage Offset Correction: v_off_ini = 138.903 uV, v_off_fin = 134.680 uV 20:10:50 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -109.277 uVs 20:10:50 Integrated Voltage Measurement... 20:10:59 Voltage Offset Correction: v_off_ini = 143.818 uV, v_off_fin = 140.673 uV 20:10:59 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = 114.590 uVs 20:10:59 Integrated Voltage Measurement... 20:11:08 Voltage Offset Correction: v_off_ini = 150.888 uV, v_off_fin = 143.657 uV 20:11:08 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -133.947 uVs 20:11:08 Integrated Voltage Measurement... 20:11:17 Voltage Offset Correction: v_off_ini = 135.259 uV, v_off_fin = 144.969 uV 20:11:17 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 138.026 uVs 20:11:21 I1X Average and RMS of Measurements VT = -119.623 +- 11.578 uVs I1X = 23.927 +- 2.314 uTm 20:11:24 Wire Position: X0 = 2.0000 mm, Y0 = 0.0000 mm 20:11:28 Integrated Voltage Measurement... 20:11:37 Voltage Offset Correction: v_off_ini = 151.106 uV, v_off_fin = 133.483 uV 20:11:37 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -184.120 uVs 20:11:37 Integrated Voltage Measurement... 20:11:46 Voltage Offset Correction: v_off_ini = 143.844 uV, v_off_fin = 158.415 uV 20:11:46 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 183.625 uVs 20:11:46 Integrated Voltage Measurement... 20:11:55 Voltage Offset Correction: v_off_ini = 152.108 uV, v_off_fin = 143.162 uV 20:11:55 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -178.937 uVs 20:11:55 Integrated Voltage Measurement... 20:12:04 Voltage Offset Correction: v_off_ini = 139.914 uV, v_off_fin = 152.618 uV 20:12:04 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 186.886 uVs 20:12:04 Integrated Voltage Measurement... 20:12:13 Voltage Offset Correction: v_off_ini = 153.666 uV, v_off_fin = 145.591 uV 20:12:13 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -180.900 uVs 20:12:13 Integrated Voltage Measurement... 20:12:22 Voltage Offset Correction: v_off_ini = 151.087 uV, v_off_fin = 165.476 uV 20:12:22 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 184.090 uVs 20:12:26 I1Y Average and RMS of Measurements VT = -183.093 +- 0.577 uVs I1Y = -36.619 +- 0.115 uTm 20:12:31 Wire Position: X0 = 2.0000 mm, Y0 = 0.0000 mm 20:12:34 Integrated Voltage Measurement... 20:12:43 Voltage Offset Correction: v_off_ini = 136.965 uV, v_off_fin = 139.754 uV 20:12:43 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -123.471 uVs 20:12:43 Integrated Voltage Measurement... 20:12:52 Voltage Offset Correction: v_off_ini = 145.154 uV, v_off_fin = 131.854 uV 20:12:52 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 120.117 uVs 20:12:52 Integrated Voltage Measurement... 20:13:01 Voltage Offset Correction: v_off_ini = 131.454 uV, v_off_fin = 132.798 uV 20:13:01 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -111.242 uVs 20:13:01 Integrated Voltage Measurement... 20:13:10 Voltage Offset Correction: v_off_ini = 136.712 uV, v_off_fin = 142.640 uV 20:13:10 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 145.626 uVs 20:13:10 Integrated Voltage Measurement... 20:13:19 Voltage Offset Correction: v_off_ini = 137.055 uV, v_off_fin = 138.680 uV 20:13:19 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -128.549 uVs 20:13:19 Integrated Voltage Measurement... 20:13:28 Voltage Offset Correction: v_off_ini = 146.963 uV, v_off_fin = 152.656 uV 20:13:28 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 144.338 uVs 20:13:30 I2X Average and RMS of Measurements VT = -128.890 +- 5.989 uVs I2X = 11.858 +- 0.551 uTm^2 20:13:36 Wire Position: X0 = 2.0000 mm, Y0 = 0.0000 mm 20:13:38 Integrated Voltage Measurement... 20:13:47 Voltage Offset Correction: v_off_ini = 143.400 uV, v_off_fin = 139.664 uV 20:13:47 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -256.796 uVs 20:13:47 Integrated Voltage Measurement... 20:13:56 Voltage Offset Correction: v_off_ini = 148.490 uV, v_off_fin = 147.840 uV 20:13:56 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 248.859 uVs 20:13:56 Integrated Voltage Measurement... 20:14:05 Voltage Offset Correction: v_off_ini = 140.125 uV, v_off_fin = 138.269 uV 20:14:05 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -269.247 uVs 20:14:05 Integrated Voltage Measurement... 20:14:14 Voltage Offset Correction: v_off_ini = 149.043 uV, v_off_fin = 144.213 uV 20:14:14 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 268.926 uVs 20:14:14 Integrated Voltage Measurement... 20:14:23 Voltage Offset Correction: v_off_ini = 125.971 uV, v_off_fin = 143.936 uV 20:14:23 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -247.650 uVs 20:14:23 Integrated Voltage Measurement... 20:14:32 Voltage Offset Correction: v_off_ini = 136.354 uV, v_off_fin = 147.311 uV 20:14:32 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 246.405 uVs 20:14:35 I2Y Average and RMS of Measurements VT = -256.314 +- 9.337 uVs I2Y = -23.581 +- 0.859 uTm^2 20:14:42 Wire Position: X0 = 0.0000 mm, Y0 = -1.0000 mm 20:14:47 Integrated Voltage Measurement... 20:14:56 Voltage Offset Correction: v_off_ini = 145.347 uV, v_off_fin = 144.875 uV 20:14:56 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -123.171 uVs 20:14:56 Integrated Voltage Measurement... 20:15:05 Voltage Offset Correction: v_off_ini = 119.378 uV, v_off_fin = 135.687 uV 20:15:05 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 119.601 uVs 20:15:05 Integrated Voltage Measurement... 20:15:14 Voltage Offset Correction: v_off_ini = 131.993 uV, v_off_fin = 131.747 uV 20:15:14 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -111.146 uVs 20:15:14 Integrated Voltage Measurement... 20:15:23 Voltage Offset Correction: v_off_ini = 138.638 uV, v_off_fin = 141.575 uV 20:15:23 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 99.160 uVs 20:15:23 Integrated Voltage Measurement... 20:15:32 Voltage Offset Correction: v_off_ini = 142.154 uV, v_off_fin = 140.836 uV 20:15:32 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -125.087 uVs 20:15:32 Integrated Voltage Measurement... 20:15:41 Voltage Offset Correction: v_off_ini = 129.642 uV, v_off_fin = 142.847 uV 20:15:41 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 113.457 uVs 20:15:45 I1X Average and RMS of Measurements VT = -115.270 +- 7.206 uVs I1X = 23.054 +- 1.442 uTm 20:15:47 Wire Position: X0 = 0.0000 mm, Y0 = -1.0000 mm 20:15:52 Integrated Voltage Measurement... 20:16:01 Voltage Offset Correction: v_off_ini = 145.380 uV, v_off_fin = 136.865 uV 20:16:01 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -177.628 uVs 20:16:01 Integrated Voltage Measurement... 20:16:10 Voltage Offset Correction: v_off_ini = 147.274 uV, v_off_fin = 154.801 uV 20:16:10 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 185.392 uVs 20:16:10 Integrated Voltage Measurement... 20:16:19 Voltage Offset Correction: v_off_ini = 143.713 uV, v_off_fin = 129.391 uV 20:16:19 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -176.078 uVs 20:16:19 Integrated Voltage Measurement... 20:16:28 Voltage Offset Correction: v_off_ini = 141.023 uV, v_off_fin = 153.548 uV 20:16:28 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 193.757 uVs 20:16:28 Integrated Voltage Measurement... 20:16:37 Voltage Offset Correction: v_off_ini = 143.549 uV, v_off_fin = 145.060 uV 20:16:37 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -175.860 uVs 20:16:37 Integrated Voltage Measurement... 20:16:46 Voltage Offset Correction: v_off_ini = 152.054 uV, v_off_fin = 170.034 uV 20:16:46 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 176.434 uVs 20:16:50 I1Y Average and RMS of Measurements VT = -180.858 +- 3.610 uVs I1Y = -36.172 +- 0.722 uTm 20:16:52 Wire Position: X0 = 0.0000 mm, Y0 = -1.0000 mm 20:16:55 Integrated Voltage Measurement... 20:17:04 Voltage Offset Correction: v_off_ini = 147.998 uV, v_off_fin = 144.969 uV 20:17:04 Integrated Voltage Measurement dXYus = -1.0001 mm, dXYds = 0.0000 mm dXY = -1.0001 mm VT = -123.063 uVs 20:17:04 Integrated Voltage Measurement... 20:17:13 Voltage Offset Correction: v_off_ini = 149.141 uV, v_off_fin = 155.415 uV 20:17:13 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 135.944 uVs 20:17:13 Integrated Voltage Measurement... 20:17:22 Voltage Offset Correction: v_off_ini = 154.475 uV, v_off_fin = 147.821 uV 20:17:22 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -130.511 uVs 20:17:22 Integrated Voltage Measurement... 20:17:31 Voltage Offset Correction: v_off_ini = 153.998 uV, v_off_fin = 145.134 uV 20:17:31 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 121.707 uVs 20:17:31 Integrated Voltage Measurement... 20:17:40 Voltage Offset Correction: v_off_ini = 132.361 uV, v_off_fin = 136.942 uV 20:17:40 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -102.724 uVs 20:17:40 Integrated Voltage Measurement... 20:17:49 Voltage Offset Correction: v_off_ini = 154.108 uV, v_off_fin = 158.859 uV 20:17:49 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 117.181 uVs 20:17:52 I2X Average and RMS of Measurements VT = -121.855 +- 8.530 uVs I2X = 11.210 +- 0.785 uTm^2 20:17:54 Wire Position: X0 = 0.0000 mm, Y0 = -1.0000 mm 20:17:57 Integrated Voltage Measurement... 20:18:06 Voltage Offset Correction: v_off_ini = 142.796 uV, v_off_fin = 138.228 uV 20:18:06 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -252.678 uVs 20:18:06 Integrated Voltage Measurement... 20:18:15 Voltage Offset Correction: v_off_ini = 157.560 uV, v_off_fin = 140.767 uV 20:18:15 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 257.696 uVs 20:18:15 Integrated Voltage Measurement... 20:18:24 Voltage Offset Correction: v_off_ini = 140.087 uV, v_off_fin = 141.271 uV 20:18:24 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -263.377 uVs 20:18:24 Integrated Voltage Measurement... 20:18:33 Voltage Offset Correction: v_off_ini = 143.025 uV, v_off_fin = 131.621 uV 20:18:33 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 258.002 uVs 20:18:33 Integrated Voltage Measurement... 20:18:42 Voltage Offset Correction: v_off_ini = 142.156 uV, v_off_fin = 152.147 uV 20:18:42 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -270.133 uVs 20:18:42 Integrated Voltage Measurement... 20:18:51 Voltage Offset Correction: v_off_ini = 147.560 uV, v_off_fin = 143.574 uV 20:18:51 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 275.374 uVs 20:18:53 I2Y Average and RMS of Measurements VT = -262.877 +- 7.337 uVs I2Y = -24.185 +- 0.675 uTm^2 20:18:57 Wire Position: X0 = 0.0000 mm, Y0 = -0.5000 mm 20:19:02 Integrated Voltage Measurement... 20:19:11 Voltage Offset Correction: v_off_ini = 170.049 uV, v_off_fin = 151.349 uV 20:19:11 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -132.860 uVs 20:19:11 Integrated Voltage Measurement... 20:19:20 Voltage Offset Correction: v_off_ini = 148.383 uV, v_off_fin = 160.505 uV 20:19:20 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 114.062 uVs 20:19:20 Integrated Voltage Measurement... 20:19:29 Voltage Offset Correction: v_off_ini = 146.512 uV, v_off_fin = 143.677 uV 20:19:29 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.4999 mm dXY = -0.4999 mm VT = -126.343 uVs 20:19:29 Integrated Voltage Measurement... 20:19:38 Voltage Offset Correction: v_off_ini = 148.929 uV, v_off_fin = 143.498 uV 20:19:38 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = 110.630 uVs 20:19:38 Integrated Voltage Measurement... 20:19:47 Voltage Offset Correction: v_off_ini = 129.031 uV, v_off_fin = 147.915 uV 20:19:47 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -106.878 uVs 20:19:47 Integrated Voltage Measurement... 20:19:56 Voltage Offset Correction: v_off_ini = 145.159 uV, v_off_fin = 143.873 uV 20:19:56 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 133.320 uVs 20:20:00 I1X Average and RMS of Measurements VT = -120.682 +- 2.072 uVs I1X = 24.138 +- 0.413 uTm 20:20:02 Wire Position: X0 = 0.0000 mm, Y0 = -0.5000 mm 20:20:07 Integrated Voltage Measurement... 20:20:16 Voltage Offset Correction: v_off_ini = 154.683 uV, v_off_fin = 139.391 uV 20:20:16 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -183.036 uVs 20:20:16 Integrated Voltage Measurement... 20:20:25 Voltage Offset Correction: v_off_ini = 147.046 uV, v_off_fin = 143.529 uV 20:20:25 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 186.500 uVs 20:20:25 Integrated Voltage Measurement... 20:20:34 Voltage Offset Correction: v_off_ini = 140.764 uV, v_off_fin = 137.324 uV 20:20:34 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -183.946 uVs 20:20:34 Integrated Voltage Measurement... 20:20:43 Voltage Offset Correction: v_off_ini = 155.793 uV, v_off_fin = 158.129 uV 20:20:43 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 162.127 uVs 20:20:43 Integrated Voltage Measurement... 20:20:52 Voltage Offset Correction: v_off_ini = 147.285 uV, v_off_fin = 141.517 uV 20:20:52 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -178.466 uVs 20:20:52 Integrated Voltage Measurement... 20:21:01 Voltage Offset Correction: v_off_ini = 149.404 uV, v_off_fin = 160.726 uV 20:21:01 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 184.555 uVs 20:21:05 I1Y Average and RMS of Measurements VT = -179.772 +- 4.945 uVs I1Y = -35.954 +- 0.989 uTm 20:21:07 Wire Position: X0 = 0.0000 mm, Y0 = -0.5000 mm 20:21:10 Integrated Voltage Measurement... 20:21:19 Voltage Offset Correction: v_off_ini = 152.046 uV, v_off_fin = 131.895 uV 20:21:19 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -152.500 uVs 20:21:19 Integrated Voltage Measurement... 20:21:28 Voltage Offset Correction: v_off_ini = 166.702 uV, v_off_fin = 165.321 uV 20:21:28 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 117.604 uVs 20:21:28 Integrated Voltage Measurement... 20:21:37 Voltage Offset Correction: v_off_ini = 162.537 uV, v_off_fin = 146.485 uV 20:21:37 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -113.086 uVs 20:21:37 Integrated Voltage Measurement... 20:21:46 Voltage Offset Correction: v_off_ini = 151.524 uV, v_off_fin = 158.942 uV 20:21:46 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 120.645 uVs 20:21:46 Integrated Voltage Measurement... 20:21:55 Voltage Offset Correction: v_off_ini = 148.547 uV, v_off_fin = 135.748 uV 20:21:55 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -137.475 uVs 20:21:55 Integrated Voltage Measurement... 20:22:04 Voltage Offset Correction: v_off_ini = 149.776 uV, v_off_fin = 154.670 uV 20:22:04 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 118.815 uVs 20:22:07 I2X Average and RMS of Measurements VT = -126.688 +- 7.496 uVs I2X = 11.655 +- 0.690 uTm^2 20:22:09 Wire Position: X0 = 0.0000 mm, Y0 = -0.5000 mm 20:22:12 Integrated Voltage Measurement... 20:22:21 Voltage Offset Correction: v_off_ini = 149.642 uV, v_off_fin = 160.296 uV 20:22:21 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -257.567 uVs 20:22:21 Integrated Voltage Measurement... 20:22:30 Voltage Offset Correction: v_off_ini = 153.709 uV, v_off_fin = 151.460 uV 20:22:30 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 257.264 uVs 20:22:30 Integrated Voltage Measurement... 20:22:39 Voltage Offset Correction: v_off_ini = 156.031 uV, v_off_fin = 162.015 uV 20:22:39 Integrated Voltage Measurement dXYus = 0.9998 mm, dXYds = 0.0000 mm dXY = 0.9998 mm VT = -253.678 uVs 20:22:39 Integrated Voltage Measurement... 20:22:48 Voltage Offset Correction: v_off_ini = 164.999 uV, v_off_fin = 153.631 uV 20:22:48 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 260.917 uVs 20:22:48 Integrated Voltage Measurement... 20:22:57 Voltage Offset Correction: v_off_ini = 133.537 uV, v_off_fin = 137.153 uV 20:22:57 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -258.261 uVs 20:22:57 Integrated Voltage Measurement... 20:23:06 Voltage Offset Correction: v_off_ini = 148.668 uV, v_off_fin = 156.226 uV 20:23:06 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 245.054 uVs 20:23:08 I2Y Average and RMS of Measurements VT = -255.457 +- 2.687 uVs I2Y = -23.503 +- 0.248 uTm^2 20:23:12 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 20:23:17 Integrated Voltage Measurement... 20:23:26 Voltage Offset Correction: v_off_ini = 150.594 uV, v_off_fin = 150.603 uV 20:23:26 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.4998 mm dXY = -0.4999 mm VT = -117.275 uVs 20:23:26 Integrated Voltage Measurement... 20:23:35 Voltage Offset Correction: v_off_ini = 153.749 uV, v_off_fin = 157.372 uV 20:23:35 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = 114.713 uVs 20:23:35 Integrated Voltage Measurement... 20:23:44 Voltage Offset Correction: v_off_ini = 145.038 uV, v_off_fin = 142.125 uV 20:23:44 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -117.457 uVs 20:23:44 Integrated Voltage Measurement... 20:23:53 Voltage Offset Correction: v_off_ini = 154.126 uV, v_off_fin = 160.743 uV 20:23:53 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 114.732 uVs 20:23:53 Integrated Voltage Measurement... 20:24:02 Voltage Offset Correction: v_off_ini = 149.797 uV, v_off_fin = 146.162 uV 20:24:02 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -132.172 uVs 20:24:02 Integrated Voltage Measurement... 20:24:11 Voltage Offset Correction: v_off_ini = 165.786 uV, v_off_fin = 162.793 uV 20:24:11 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 124.892 uVs 20:24:15 I1X Average and RMS of Measurements VT = -120.207 +- 5.887 uVs I1X = 24.043 +- 1.176 uTm 20:24:18 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 20:24:22 Integrated Voltage Measurement... 20:24:31 Voltage Offset Correction: v_off_ini = 166.882 uV, v_off_fin = 157.149 uV 20:24:31 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -170.220 uVs 20:24:31 Integrated Voltage Measurement... 20:24:40 Voltage Offset Correction: v_off_ini = 155.895 uV, v_off_fin = 158.846 uV 20:24:40 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 182.975 uVs 20:24:40 Integrated Voltage Measurement... 20:24:49 Voltage Offset Correction: v_off_ini = 158.760 uV, v_off_fin = 162.031 uV 20:24:49 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -190.248 uVs 20:24:49 Integrated Voltage Measurement... 20:24:58 Voltage Offset Correction: v_off_ini = 163.810 uV, v_off_fin = 172.728 uV 20:24:58 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 180.877 uVs 20:24:58 Integrated Voltage Measurement... 20:25:07 Voltage Offset Correction: v_off_ini = 166.686 uV, v_off_fin = 147.360 uV 20:25:07 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -178.085 uVs 20:25:07 Integrated Voltage Measurement... 20:25:16 Voltage Offset Correction: v_off_ini = 173.536 uV, v_off_fin = 175.732 uV 20:25:16 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 176.923 uVs 20:25:20 I1Y Average and RMS of Measurements VT = -179.888 +- 4.029 uVs I1Y = -35.978 +- 0.806 uTm 20:25:23 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 20:25:25 Integrated Voltage Measurement... 20:25:34 Voltage Offset Correction: v_off_ini = 172.797 uV, v_off_fin = 157.418 uV 20:25:34 Integrated Voltage Measurement dXYus = -1.0001 mm, dXYds = 0.0000 mm dXY = -1.0001 mm VT = -117.985 uVs 20:25:34 Integrated Voltage Measurement... 20:25:43 Voltage Offset Correction: v_off_ini = 166.636 uV, v_off_fin = 171.420 uV 20:25:43 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 145.027 uVs 20:25:43 Integrated Voltage Measurement... 20:25:52 Voltage Offset Correction: v_off_ini = 167.615 uV, v_off_fin = 159.944 uV 20:25:52 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -135.074 uVs 20:25:52 Integrated Voltage Measurement... 20:26:01 Voltage Offset Correction: v_off_ini = 165.910 uV, v_off_fin = 171.013 uV 20:26:01 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 112.509 uVs 20:26:01 Integrated Voltage Measurement... 20:26:10 Voltage Offset Correction: v_off_ini = 167.079 uV, v_off_fin = 154.373 uV 20:26:10 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -143.333 uVs 20:26:10 Integrated Voltage Measurement... 20:26:19 Voltage Offset Correction: v_off_ini = 158.902 uV, v_off_fin = 173.108 uV 20:26:19 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 122.229 uVs 20:26:22 I2X Average and RMS of Measurements VT = -129.360 +- 3.971 uVs I2X = 11.901 +- 0.365 uTm^2 20:26:24 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 20:26:27 Integrated Voltage Measurement... 20:26:36 Voltage Offset Correction: v_off_ini = 163.716 uV, v_off_fin = 165.169 uV 20:26:36 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -261.215 uVs 20:26:36 Integrated Voltage Measurement... 20:26:45 Voltage Offset Correction: v_off_ini = 177.386 uV, v_off_fin = 172.797 uV 20:26:45 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 266.137 uVs 20:26:45 Integrated Voltage Measurement... 20:26:54 Voltage Offset Correction: v_off_ini = 183.779 uV, v_off_fin = 180.250 uV 20:26:54 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -268.865 uVs 20:26:54 Integrated Voltage Measurement... 20:27:03 Voltage Offset Correction: v_off_ini = 181.169 uV, v_off_fin = 184.848 uV 20:27:03 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 264.079 uVs 20:27:03 Integrated Voltage Measurement... 20:27:12 Voltage Offset Correction: v_off_ini = 182.227 uV, v_off_fin = 176.758 uV 20:27:12 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = -266.851 uVs 20:27:12 Integrated Voltage Measurement... 20:27:21 Voltage Offset Correction: v_off_ini = 175.058 uV, v_off_fin = 195.590 uV 20:27:21 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 266.387 uVs 20:27:24 I2Y Average and RMS of Measurements VT = -265.589 +- 1.354 uVs I2Y = -24.435 +- 0.125 uTm^2 20:27:28 Wire Position: X0 = 0.0000 mm, Y0 = 0.5000 mm 20:27:32 Integrated Voltage Measurement... 20:27:41 Voltage Offset Correction: v_off_ini = 168.346 uV, v_off_fin = 154.261 uV 20:27:41 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -129.477 uVs 20:27:41 Integrated Voltage Measurement... 20:27:50 Voltage Offset Correction: v_off_ini = 151.232 uV, v_off_fin = 164.018 uV 20:27:50 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 93.855 uVs 20:27:50 Integrated Voltage Measurement... 20:27:59 Voltage Offset Correction: v_off_ini = 165.527 uV, v_off_fin = 156.184 uV 20:27:59 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -120.208 uVs 20:27:59 Integrated Voltage Measurement... 20:28:08 Voltage Offset Correction: v_off_ini = 160.037 uV, v_off_fin = 168.674 uV 20:28:08 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 95.563 uVs 20:28:08 Integrated Voltage Measurement... 20:28:17 Voltage Offset Correction: v_off_ini = 157.374 uV, v_off_fin = 160.558 uV 20:28:17 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -119.182 uVs 20:28:17 Integrated Voltage Measurement... 20:28:26 Voltage Offset Correction: v_off_ini = 180.161 uV, v_off_fin = 173.907 uV 20:28:26 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = 87.030 uVs 20:28:30 I1X Average and RMS of Measurements VT = -107.552 +- 3.502 uVs I1X = 21.512 +- 0.700 uTm 20:28:33 Wire Position: X0 = 0.0000 mm, Y0 = 0.5000 mm 20:28:37 Integrated Voltage Measurement... 20:28:46 Voltage Offset Correction: v_off_ini = 168.482 uV, v_off_fin = 171.242 uV 20:28:46 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -176.331 uVs 20:28:46 Integrated Voltage Measurement... 20:28:55 Voltage Offset Correction: v_off_ini = 184.209 uV, v_off_fin = 194.478 uV 20:28:55 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 161.969 uVs 20:28:55 Integrated Voltage Measurement... 20:29:04 Voltage Offset Correction: v_off_ini = 172.017 uV, v_off_fin = 166.368 uV 20:29:04 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -175.831 uVs 20:29:04 Integrated Voltage Measurement... 20:29:13 Voltage Offset Correction: v_off_ini = 170.926 uV, v_off_fin = 188.960 uV 20:29:13 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 177.471 uVs 20:29:13 Integrated Voltage Measurement... 20:29:22 Voltage Offset Correction: v_off_ini = 176.884 uV, v_off_fin = 176.990 uV 20:29:22 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -167.348 uVs 20:29:22 Integrated Voltage Measurement... 20:29:31 Voltage Offset Correction: v_off_ini = 189.038 uV, v_off_fin = 182.650 uV 20:29:31 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 178.521 uVs 20:29:35 I1Y Average and RMS of Measurements VT = -172.912 +- 3.062 uVs I1Y = -34.580 +- 0.612 uTm 20:29:38 Wire Position: X0 = 0.0000 mm, Y0 = 0.5000 mm 20:29:40 Integrated Voltage Measurement... 20:29:49 Voltage Offset Correction: v_off_ini = 200.069 uV, v_off_fin = 185.305 uV 20:29:49 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -134.743 uVs 20:29:49 Integrated Voltage Measurement... 20:29:58 Voltage Offset Correction: v_off_ini = 188.756 uV, v_off_fin = 200.809 uV 20:29:58 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 122.970 uVs 20:29:58 Integrated Voltage Measurement... 20:30:07 Voltage Offset Correction: v_off_ini = 184.631 uV, v_off_fin = 175.182 uV 20:30:07 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -122.628 uVs 20:30:07 Integrated Voltage Measurement... 20:30:16 Voltage Offset Correction: v_off_ini = 181.908 uV, v_off_fin = 194.904 uV 20:30:16 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 123.263 uVs 20:30:16 Integrated Voltage Measurement... 20:30:25 Voltage Offset Correction: v_off_ini = 197.483 uV, v_off_fin = 187.224 uV 20:30:25 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -124.577 uVs 20:30:25 Integrated Voltage Measurement... 20:30:34 Voltage Offset Correction: v_off_ini = 195.314 uV, v_off_fin = 199.996 uV 20:30:35 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 114.835 uVs 20:30:37 I2X Average and RMS of Measurements VT = -123.836 +- 3.789 uVs I2X = 11.393 +- 0.349 uTm^2 20:30:39 Wire Position: X0 = 0.0000 mm, Y0 = 0.5000 mm 20:30:42 Integrated Voltage Measurement... 20:30:51 Voltage Offset Correction: v_off_ini = 186.835 uV, v_off_fin = 188.074 uV 20:30:51 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -245.862 uVs 20:30:51 Integrated Voltage Measurement... 20:31:00 Voltage Offset Correction: v_off_ini = 179.504 uV, v_off_fin = 191.313 uV 20:31:00 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 277.157 uVs 20:31:00 Integrated Voltage Measurement... 20:31:09 Voltage Offset Correction: v_off_ini = 197.974 uV, v_off_fin = 179.433 uV 20:31:09 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -257.192 uVs 20:31:09 Integrated Voltage Measurement... 20:31:18 Voltage Offset Correction: v_off_ini = 182.181 uV, v_off_fin = 195.523 uV 20:31:18 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 257.211 uVs 20:31:18 Integrated Voltage Measurement... 20:31:27 Voltage Offset Correction: v_off_ini = 196.126 uV, v_off_fin = 191.480 uV 20:31:27 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -265.207 uVs 20:31:27 Integrated Voltage Measurement... 20:31:36 Voltage Offset Correction: v_off_ini = 188.073 uV, v_off_fin = 176.185 uV 20:31:36 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 268.690 uVs 20:31:39 I2Y Average and RMS of Measurements VT = -261.887 +- 3.988 uVs I2Y = -24.094 +- 0.367 uTm^2 20:31:43 Wire Position: X0 = 0.0000 mm, Y0 = 1.0000 mm 20:31:47 Integrated Voltage Measurement... 20:31:56 Voltage Offset Correction: v_off_ini = 181.481 uV, v_off_fin = 172.663 uV 20:31:56 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -98.550 uVs 20:31:56 Integrated Voltage Measurement... 20:32:05 Voltage Offset Correction: v_off_ini = 184.871 uV, v_off_fin = 190.664 uV 20:32:05 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 107.217 uVs 20:32:05 Integrated Voltage Measurement... 20:32:14 Voltage Offset Correction: v_off_ini = 166.842 uV, v_off_fin = 180.007 uV 20:32:14 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -111.254 uVs 20:32:14 Integrated Voltage Measurement... 20:32:23 Voltage Offset Correction: v_off_ini = 188.698 uV, v_off_fin = 180.250 uV 20:32:23 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = 111.453 uVs 20:32:23 Integrated Voltage Measurement... 20:32:32 Voltage Offset Correction: v_off_ini = 166.441 uV, v_off_fin = 175.095 uV 20:32:32 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -99.520 uVs 20:32:32 Integrated Voltage Measurement... 20:32:41 Voltage Offset Correction: v_off_ini = 192.306 uV, v_off_fin = 189.672 uV 20:32:41 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 101.547 uVs 20:32:45 I1X Average and RMS of Measurements VT = -104.924 +- 4.647 uVs I1X = 20.985 +- 0.930 uTm 20:32:48 Wire Position: X0 = 0.0000 mm, Y0 = 1.0000 mm 20:32:52 Integrated Voltage Measurement... 20:33:01 Voltage Offset Correction: v_off_ini = 170.392 uV, v_off_fin = 151.211 uV 20:33:01 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -179.167 uVs 20:33:01 Integrated Voltage Measurement... 20:33:10 Voltage Offset Correction: v_off_ini = 171.142 uV, v_off_fin = 182.772 uV 20:33:10 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 171.286 uVs 20:33:10 Integrated Voltage Measurement... 20:33:19 Voltage Offset Correction: v_off_ini = 160.172 uV, v_off_fin = 149.632 uV 20:33:19 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -175.442 uVs 20:33:19 Integrated Voltage Measurement... 20:33:28 Voltage Offset Correction: v_off_ini = 160.838 uV, v_off_fin = 176.095 uV 20:33:28 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 165.976 uVs 20:33:28 Integrated Voltage Measurement... 20:33:37 Voltage Offset Correction: v_off_ini = 165.990 uV, v_off_fin = 151.699 uV 20:33:37 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -178.054 uVs 20:33:37 Integrated Voltage Measurement... 20:33:46 Voltage Offset Correction: v_off_ini = 170.714 uV, v_off_fin = 172.118 uV 20:33:46 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 180.577 uVs 20:33:50 I1Y Average and RMS of Measurements VT = -175.084 +- 3.515 uVs I1Y = -35.016 +- 0.702 uTm 20:33:53 Wire Position: X0 = 0.0000 mm, Y0 = 1.0000 mm 20:33:55 Integrated Voltage Measurement... 20:34:04 Voltage Offset Correction: v_off_ini = 163.084 uV, v_off_fin = 148.468 uV 20:34:04 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -120.162 uVs 20:34:04 Integrated Voltage Measurement... 20:34:13 Voltage Offset Correction: v_off_ini = 168.912 uV, v_off_fin = 176.585 uV 20:34:14 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 117.126 uVs 20:34:14 Integrated Voltage Measurement... 20:34:23 Voltage Offset Correction: v_off_ini = 169.854 uV, v_off_fin = 153.914 uV 20:34:23 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -108.824 uVs 20:34:23 Integrated Voltage Measurement... 20:34:32 Voltage Offset Correction: v_off_ini = 158.037 uV, v_off_fin = 170.946 uV 20:34:32 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 105.461 uVs 20:34:32 Integrated Voltage Measurement... 20:34:41 Voltage Offset Correction: v_off_ini = 172.032 uV, v_off_fin = 157.117 uV 20:34:41 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -126.984 uVs 20:34:41 Integrated Voltage Measurement... 20:34:50 Voltage Offset Correction: v_off_ini = 161.249 uV, v_off_fin = 170.373 uV 20:34:50 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 123.625 uVs 20:34:52 I2X Average and RMS of Measurements VT = -117.030 +- 7.502 uVs I2X = 10.767 +- 0.690 uTm^2 20:34:55 Wire Position: X0 = 0.0000 mm, Y0 = 1.0000 mm 20:34:57 Integrated Voltage Measurement... 20:35:06 Voltage Offset Correction: v_off_ini = 154.482 uV, v_off_fin = 157.560 uV 20:35:06 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -267.374 uVs 20:35:06 Integrated Voltage Measurement... 20:35:15 Voltage Offset Correction: v_off_ini = 158.174 uV, v_off_fin = 160.429 uV 20:35:15 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 275.481 uVs 20:35:15 Integrated Voltage Measurement... 20:35:24 Voltage Offset Correction: v_off_ini = 166.800 uV, v_off_fin = 157.018 uV 20:35:24 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -261.195 uVs 20:35:24 Integrated Voltage Measurement... 20:35:33 Voltage Offset Correction: v_off_ini = 154.285 uV, v_off_fin = 162.161 uV 20:35:33 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 258.275 uVs 20:35:33 Integrated Voltage Measurement... 20:35:42 Voltage Offset Correction: v_off_ini = 157.010 uV, v_off_fin = 151.506 uV 20:35:42 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -273.382 uVs 20:35:42 Integrated Voltage Measurement... 20:35:51 Voltage Offset Correction: v_off_ini = 167.849 uV, v_off_fin = 158.765 uV 20:35:51 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 267.770 uVs 20:35:54 I2Y Average and RMS of Measurements VT = -267.246 +- 5.323 uVs I2Y = -24.587 +- 0.490 uTm^2 20:36:11 Wire Position: X0 = -2.0000 mm, Y0 = 0.0000 mm 20:36:15 Integrated Voltage Measurement... 20:36:24 Voltage Offset Correction: v_off_ini = 155.636 uV, v_off_fin = 146.634 uV 20:36:24 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -142.349 uVs 20:36:24 Integrated Voltage Measurement... 20:36:33 Voltage Offset Correction: v_off_ini = 155.086 uV, v_off_fin = 154.516 uV 20:36:33 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 125.665 uVs 20:36:33 Integrated Voltage Measurement... 20:36:42 Voltage Offset Correction: v_off_ini = 154.267 uV, v_off_fin = 136.473 uV 20:36:42 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -100.854 uVs 20:36:42 Integrated Voltage Measurement... 20:36:51 Voltage Offset Correction: v_off_ini = 145.808 uV, v_off_fin = 146.578 uV 20:36:51 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 113.594 uVs 20:36:51 Integrated Voltage Measurement... 20:37:00 Voltage Offset Correction: v_off_ini = 143.336 uV, v_off_fin = 127.470 uV 20:37:00 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -130.123 uVs 20:37:00 Integrated Voltage Measurement... 20:37:09 Voltage Offset Correction: v_off_ini = 123.325 uV, v_off_fin = 143.617 uV 20:37:09 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 140.204 uVs 20:37:13 I1X Average and RMS of Measurements VT = -125.465 +- 12.907 uVs I1X = 25.093 +- 2.581 uTm 20:37:16 Wire Position: X0 = -2.0000 mm, Y0 = 0.0000 mm 20:37:20 Integrated Voltage Measurement... 20:37:29 Voltage Offset Correction: v_off_ini = 140.999 uV, v_off_fin = 108.512 uV 20:37:29 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -179.140 uVs 20:37:29 Integrated Voltage Measurement... 20:37:38 Voltage Offset Correction: v_off_ini = 141.750 uV, v_off_fin = 160.759 uV 20:37:38 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 173.015 uVs 20:37:38 Integrated Voltage Measurement... 20:37:47 Voltage Offset Correction: v_off_ini = 139.315 uV, v_off_fin = 135.316 uV 20:37:47 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -192.257 uVs 20:37:47 Integrated Voltage Measurement... 20:37:56 Voltage Offset Correction: v_off_ini = 125.080 uV, v_off_fin = 141.332 uV 20:37:56 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 189.985 uVs 20:37:56 Integrated Voltage Measurement... 20:38:05 Voltage Offset Correction: v_off_ini = 143.099 uV, v_off_fin = 130.823 uV 20:38:05 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -181.060 uVs 20:38:05 Integrated Voltage Measurement... 20:38:14 Voltage Offset Correction: v_off_ini = 126.976 uV, v_off_fin = 133.577 uV 20:38:14 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 188.661 uVs 20:38:18 I1Y Average and RMS of Measurements VT = -184.020 +- 6.170 uVs I1Y = -36.804 +- 1.234 uTm 20:38:24 Wire Position: X0 = -2.0000 mm, Y0 = 0.0000 mm 20:38:26 Integrated Voltage Measurement... 20:38:35 Voltage Offset Correction: v_off_ini = 126.267 uV, v_off_fin = 115.572 uV 20:38:35 Integrated Voltage Measurement dXYus = -1.0001 mm, dXYds = 0.0000 mm dXY = -1.0001 mm VT = -133.552 uVs 20:38:35 Integrated Voltage Measurement... 20:38:44 Voltage Offset Correction: v_off_ini = 121.320 uV, v_off_fin = 123.760 uV 20:38:44 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 108.179 uVs 20:38:44 Integrated Voltage Measurement... 20:38:53 Voltage Offset Correction: v_off_ini = 115.302 uV, v_off_fin = 99.501 uV 20:38:53 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -150.564 uVs 20:38:53 Integrated Voltage Measurement... 20:39:02 Voltage Offset Correction: v_off_ini = 113.322 uV, v_off_fin = 131.061 uV 20:39:02 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 131.958 uVs 20:39:02 Integrated Voltage Measurement... 20:39:11 Voltage Offset Correction: v_off_ini = 129.568 uV, v_off_fin = 115.448 uV 20:39:11 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -140.635 uVs 20:39:11 Integrated Voltage Measurement... 20:39:20 Voltage Offset Correction: v_off_ini = 135.183 uV, v_off_fin = 133.609 uV 20:39:20 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 125.325 uVs 20:39:23 I2X Average and RMS of Measurements VT = -131.702 +- 8.375 uVs I2X = 12.117 +- 0.771 uTm^2 20:39:28 Wire Position: X0 = -2.0000 mm, Y0 = 0.0000 mm 20:39:30 Integrated Voltage Measurement... 20:39:39 Voltage Offset Correction: v_off_ini = 134.731 uV, v_off_fin = 117.729 uV 20:39:39 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -262.761 uVs 20:39:40 Integrated Voltage Measurement... 20:39:48 Voltage Offset Correction: v_off_ini = 138.812 uV, v_off_fin = 137.189 uV 20:39:49 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 244.654 uVs 20:39:49 Integrated Voltage Measurement... 20:39:58 Voltage Offset Correction: v_off_ini = 128.087 uV, v_off_fin = 124.532 uV 20:39:58 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = -249.319 uVs 20:39:58 Integrated Voltage Measurement... 20:40:07 Voltage Offset Correction: v_off_ini = 132.209 uV, v_off_fin = 141.565 uV 20:40:07 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 245.893 uVs 20:40:07 Integrated Voltage Measurement... 20:40:16 Voltage Offset Correction: v_off_ini = 138.178 uV, v_off_fin = 126.035 uV 20:40:16 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -266.982 uVs 20:40:16 Integrated Voltage Measurement... 20:40:25 Voltage Offset Correction: v_off_ini = 132.467 uV, v_off_fin = 127.942 uV 20:40:25 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 254.231 uVs 20:40:27 I2Y Average and RMS of Measurements VT = -253.973 +- 5.311 uVs I2Y = -23.366 +- 0.488 uTm^2 20:40:32 Wire Position: X0 = -1.0000 mm, Y0 = 0.0000 mm 20:40:36 Integrated Voltage Measurement... 20:40:45 Voltage Offset Correction: v_off_ini = 110.395 uV, v_off_fin = 116.771 uV 20:40:45 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -119.416 uVs 20:40:45 Integrated Voltage Measurement... 20:40:54 Voltage Offset Correction: v_off_ini = 122.406 uV, v_off_fin = 107.333 uV 20:40:54 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 130.156 uVs 20:40:54 Integrated Voltage Measurement... 20:41:03 Voltage Offset Correction: v_off_ini = 112.013 uV, v_off_fin = 97.808 uV 20:41:03 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -114.427 uVs 20:41:03 Integrated Voltage Measurement... 20:41:12 Voltage Offset Correction: v_off_ini = 99.751 uV, v_off_fin = 103.801 uV 20:41:12 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 123.372 uVs 20:41:12 Integrated Voltage Measurement... 20:41:21 Voltage Offset Correction: v_off_ini = 103.820 uV, v_off_fin = 103.429 uV 20:41:21 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -106.490 uVs 20:41:21 Integrated Voltage Measurement... 20:41:30 Voltage Offset Correction: v_off_ini = 107.285 uV, v_off_fin = 100.858 uV 20:41:30 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 131.065 uVs 20:41:34 I1X Average and RMS of Measurements VT = -120.821 +- 2.804 uVs I1X = 24.164 +- 0.561 uTm 20:41:37 Wire Position: X0 = -1.0000 mm, Y0 = 0.0000 mm 20:41:41 Integrated Voltage Measurement... 20:41:50 Voltage Offset Correction: v_off_ini = 95.951 uV, v_off_fin = 96.669 uV 20:41:50 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -189.140 uVs 20:41:50 Integrated Voltage Measurement... 20:41:59 Voltage Offset Correction: v_off_ini = 110.310 uV, v_off_fin = 108.762 uV 20:41:59 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 201.521 uVs 20:41:59 Integrated Voltage Measurement... 20:42:08 Voltage Offset Correction: v_off_ini = 109.664 uV, v_off_fin = 95.119 uV 20:42:08 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -193.530 uVs 20:42:08 Integrated Voltage Measurement... 20:42:17 Voltage Offset Correction: v_off_ini = 109.449 uV, v_off_fin = 108.672 uV 20:42:17 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 193.057 uVs 20:42:17 Integrated Voltage Measurement... 20:42:26 Voltage Offset Correction: v_off_ini = 102.295 uV, v_off_fin = 97.972 uV 20:42:26 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -185.933 uVs 20:42:26 Integrated Voltage Measurement... 20:42:35 Voltage Offset Correction: v_off_ini = 106.226 uV, v_off_fin = 96.465 uV 20:42:35 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 182.972 uVs 20:42:39 I1Y Average and RMS of Measurements VT = -191.025 +- 4.722 uVs I1Y = -38.205 +- 0.944 uTm 20:42:44 Wire Position: X0 = -1.0000 mm, Y0 = 0.0000 mm 20:42:47 Integrated Voltage Measurement... 20:42:56 Voltage Offset Correction: v_off_ini = 106.982 uV, v_off_fin = 86.599 uV 20:42:56 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -156.942 uVs 20:42:56 Integrated Voltage Measurement... 20:43:05 Voltage Offset Correction: v_off_ini = 88.261 uV, v_off_fin = 107.447 uV 20:43:05 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 123.543 uVs 20:43:05 Integrated Voltage Measurement... 20:43:14 Voltage Offset Correction: v_off_ini = 107.794 uV, v_off_fin = 93.023 uV 20:43:14 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -143.731 uVs 20:43:14 Integrated Voltage Measurement... 20:43:23 Voltage Offset Correction: v_off_ini = 105.184 uV, v_off_fin = 116.805 uV 20:43:23 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 130.894 uVs 20:43:23 Integrated Voltage Measurement... 20:43:32 Voltage Offset Correction: v_off_ini = 106.819 uV, v_off_fin = 103.152 uV 20:43:32 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -141.647 uVs 20:43:32 Integrated Voltage Measurement... 20:43:41 Voltage Offset Correction: v_off_ini = 118.145 uV, v_off_fin = 120.625 uV 20:43:41 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 141.710 uVs 20:43:44 I2X Average and RMS of Measurements VT = -139.744 +- 1.817 uVs I2X = 12.857 +- 0.167 uTm^2 20:43:49 Wire Position: X0 = -1.0000 mm, Y0 = 0.0000 mm 20:43:51 Integrated Voltage Measurement... 20:44:00 Voltage Offset Correction: v_off_ini = 108.698 uV, v_off_fin = 117.567 uV 20:44:00 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -251.358 uVs 20:44:00 Integrated Voltage Measurement... 20:44:09 Voltage Offset Correction: v_off_ini = 115.194 uV, v_off_fin = 116.464 uV 20:44:09 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 256.004 uVs 20:44:10 Integrated Voltage Measurement... 20:44:18 Voltage Offset Correction: v_off_ini = 123.383 uV, v_off_fin = 119.092 uV 20:44:19 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -257.024 uVs 20:44:19 Integrated Voltage Measurement... 20:44:28 Voltage Offset Correction: v_off_ini = 105.515 uV, v_off_fin = 98.280 uV 20:44:28 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 262.756 uVs 20:44:28 Integrated Voltage Measurement... 20:44:37 Voltage Offset Correction: v_off_ini = 93.052 uV, v_off_fin = 101.724 uV 20:44:37 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -254.584 uVs 20:44:37 Integrated Voltage Measurement... 20:44:46 Voltage Offset Correction: v_off_ini = 107.892 uV, v_off_fin = 95.367 uV 20:44:46 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 274.410 uVs 20:44:48 I2Y Average and RMS of Measurements VT = -259.356 +- 4.432 uVs I2Y = -23.861 +- 0.408 uTm^2 20:44:53 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 20:44:57 Integrated Voltage Measurement... 20:45:06 Voltage Offset Correction: v_off_ini = 103.831 uV, v_off_fin = 99.314 uV 20:45:06 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -116.590 uVs 20:45:06 Integrated Voltage Measurement... 20:45:15 Voltage Offset Correction: v_off_ini = 109.959 uV, v_off_fin = 117.232 uV 20:45:15 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 133.615 uVs 20:45:15 Integrated Voltage Measurement... 20:45:24 Voltage Offset Correction: v_off_ini = 105.532 uV, v_off_fin = 108.282 uV 20:45:24 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -121.980 uVs 20:45:24 Integrated Voltage Measurement... 20:45:33 Voltage Offset Correction: v_off_ini = 110.148 uV, v_off_fin = 115.965 uV 20:45:33 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 121.393 uVs 20:45:33 Integrated Voltage Measurement... 20:45:42 Voltage Offset Correction: v_off_ini = 97.611 uV, v_off_fin = 92.947 uV 20:45:42 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -137.883 uVs 20:45:42 Integrated Voltage Measurement... 20:45:51 Voltage Offset Correction: v_off_ini = 88.893 uV, v_off_fin = 106.963 uV 20:45:51 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 137.443 uVs 20:45:55 I1X Average and RMS of Measurements VT = -128.151 +- 6.869 uVs I1X = 25.630 +- 1.374 uTm 20:45:58 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 20:46:02 Integrated Voltage Measurement... 20:46:11 Voltage Offset Correction: v_off_ini = 107.484 uV, v_off_fin = 97.049 uV 20:46:11 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -182.244 uVs 20:46:11 Integrated Voltage Measurement... 20:46:20 Voltage Offset Correction: v_off_ini = 97.727 uV, v_off_fin = 118.370 uV 20:46:20 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 179.290 uVs 20:46:20 Integrated Voltage Measurement... 20:46:29 Voltage Offset Correction: v_off_ini = 97.945 uV, v_off_fin = 92.324 uV 20:46:29 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -171.195 uVs 20:46:29 Integrated Voltage Measurement... 20:46:38 Voltage Offset Correction: v_off_ini = 102.076 uV, v_off_fin = 107.380 uV 20:46:38 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 186.057 uVs 20:46:38 Integrated Voltage Measurement... 20:46:47 Voltage Offset Correction: v_off_ini = 104.322 uV, v_off_fin = 95.398 uV 20:46:47 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -179.122 uVs 20:46:47 Integrated Voltage Measurement... 20:46:56 Voltage Offset Correction: v_off_ini = 105.934 uV, v_off_fin = 118.837 uV 20:46:56 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 174.056 uVs 20:47:00 I1Y Average and RMS of Measurements VT = -178.661 +- 1.706 uVs I1Y = -35.732 +- 0.341 uTm 20:47:05 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 20:47:08 Integrated Voltage Measurement... 20:47:17 Voltage Offset Correction: v_off_ini = 101.891 uV, v_off_fin = 100.842 uV 20:47:17 Integrated Voltage Measurement dXYus = -1.0001 mm, dXYds = 0.0000 mm dXY = -1.0001 mm VT = -115.160 uVs 20:47:17 Integrated Voltage Measurement... 20:47:26 Voltage Offset Correction: v_off_ini = 108.109 uV, v_off_fin = 117.415 uV 20:47:26 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 114.562 uVs 20:47:26 Integrated Voltage Measurement... 20:47:35 Voltage Offset Correction: v_off_ini = 103.997 uV, v_off_fin = 102.117 uV 20:47:35 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -139.643 uVs 20:47:35 Integrated Voltage Measurement... 20:47:44 Voltage Offset Correction: v_off_ini = 105.808 uV, v_off_fin = 106.992 uV 20:47:44 Integrated Voltage Measurement dXYus = 0.9998 mm, dXYds = 0.0000 mm dXY = 0.9998 mm VT = 137.483 uVs 20:47:44 Integrated Voltage Measurement... 20:47:53 Voltage Offset Correction: v_off_ini = 99.698 uV, v_off_fin = 88.770 uV 20:47:53 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -123.624 uVs 20:47:53 Integrated Voltage Measurement... 20:48:02 Voltage Offset Correction: v_off_ini = 107.622 uV, v_off_fin = 116.520 uV 20:48:02 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 127.879 uVs 20:48:05 I2X Average and RMS of Measurements VT = -126.392 +- 9.687 uVs I2X = 11.629 +- 0.892 uTm^2 20:48:10 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 20:48:12 Integrated Voltage Measurement... 20:48:21 Voltage Offset Correction: v_off_ini = 91.700 uV, v_off_fin = 89.988 uV 20:48:21 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -254.869 uVs 20:48:21 Integrated Voltage Measurement... 20:48:30 Voltage Offset Correction: v_off_ini = 81.338 uV, v_off_fin = 102.235 uV 20:48:30 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 263.402 uVs 20:48:30 Integrated Voltage Measurement... 20:48:39 Voltage Offset Correction: v_off_ini = 97.067 uV, v_off_fin = 102.632 uV 20:48:39 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -265.734 uVs 20:48:39 Integrated Voltage Measurement... 20:48:48 Voltage Offset Correction: v_off_ini = 90.725 uV, v_off_fin = 95.367 uV 20:48:49 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 253.758 uVs 20:48:49 Integrated Voltage Measurement... 20:48:58 Voltage Offset Correction: v_off_ini = 92.230 uV, v_off_fin = 98.157 uV 20:48:58 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -254.754 uVs 20:48:58 Integrated Voltage Measurement... 20:49:07 Voltage Offset Correction: v_off_ini = 84.565 uV, v_off_fin = 90.946 uV 20:49:07 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 253.751 uVs 20:49:09 I2Y Average and RMS of Measurements VT = -257.711 +- 2.458 uVs I2Y = -23.709 +- 0.226 uTm^2 20:49:14 Wire Position: X0 = 1.0000 mm, Y0 = 0.0000 mm 20:49:18 Integrated Voltage Measurement... 20:49:27 Voltage Offset Correction: v_off_ini = 114.897 uV, v_off_fin = 100.238 uV 20:49:27 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -137.962 uVs 20:49:27 Integrated Voltage Measurement... 20:49:36 Voltage Offset Correction: v_off_ini = 118.637 uV, v_off_fin = 117.279 uV 20:49:36 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 133.054 uVs 20:49:36 Integrated Voltage Measurement... 20:49:45 Voltage Offset Correction: v_off_ini = 123.682 uV, v_off_fin = 106.899 uV 20:49:45 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -131.183 uVs 20:49:45 Integrated Voltage Measurement... 20:49:54 Voltage Offset Correction: v_off_ini = 122.461 uV, v_off_fin = 130.399 uV 20:49:54 Integrated Voltage Measurement dXYus = 0.4998 mm, dXYds = 0.4998 mm dXY = 0.4998 mm VT = 114.055 uVs 20:49:54 Integrated Voltage Measurement... 20:50:03 Voltage Offset Correction: v_off_ini = 124.448 uV, v_off_fin = 132.705 uV 20:50:03 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -136.870 uVs 20:50:03 Integrated Voltage Measurement... 20:50:12 Voltage Offset Correction: v_off_ini = 131.638 uV, v_off_fin = 129.439 uV 20:50:12 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.4999 mm dXY = 0.4999 mm VT = 121.708 uVs 20:50:16 I1X Average and RMS of Measurements VT = -129.139 +- 5.263 uVs I1X = 25.831 +- 1.051 uTm 20:50:19 Wire Position: X0 = 1.0000 mm, Y0 = 0.0000 mm 20:50:23 Integrated Voltage Measurement... 20:50:32 Voltage Offset Correction: v_off_ini = 127.799 uV, v_off_fin = 126.070 uV 20:50:32 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -194.499 uVs 20:50:32 Integrated Voltage Measurement... 20:50:41 Voltage Offset Correction: v_off_ini = 131.706 uV, v_off_fin = 147.209 uV 20:50:41 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 182.215 uVs 20:50:41 Integrated Voltage Measurement... 20:50:50 Voltage Offset Correction: v_off_ini = 128.196 uV, v_off_fin = 124.768 uV 20:50:50 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -175.390 uVs 20:50:50 Integrated Voltage Measurement... 20:50:59 Voltage Offset Correction: v_off_ini = 123.877 uV, v_off_fin = 137.270 uV 20:50:59 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 208.375 uVs 20:50:59 Integrated Voltage Measurement... 20:51:08 Voltage Offset Correction: v_off_ini = 125.899 uV, v_off_fin = 118.472 uV 20:51:08 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -175.563 uVs 20:51:08 Integrated Voltage Measurement... 20:51:17 Voltage Offset Correction: v_off_ini = 127.533 uV, v_off_fin = 139.236 uV 20:51:17 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 175.333 uVs 20:51:21 I1Y Average and RMS of Measurements VT = -185.229 +- 7.065 uVs I1Y = -37.046 +- 1.413 uTm 20:51:26 Wire Position: X0 = 1.0000 mm, Y0 = 0.0000 mm 20:51:29 Integrated Voltage Measurement... 20:51:38 Voltage Offset Correction: v_off_ini = 129.379 uV, v_off_fin = 118.654 uV 20:51:38 Integrated Voltage Measurement dXYus = -1.0001 mm, dXYds = 0.0000 mm dXY = -1.0001 mm VT = -136.819 uVs 20:51:38 Integrated Voltage Measurement... 20:51:47 Voltage Offset Correction: v_off_ini = 138.469 uV, v_off_fin = 144.458 uV 20:51:47 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 111.806 uVs 20:51:47 Integrated Voltage Measurement... 20:51:56 Voltage Offset Correction: v_off_ini = 129.547 uV, v_off_fin = 134.328 uV 20:51:56 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -125.908 uVs 20:51:56 Integrated Voltage Measurement... 20:52:05 Voltage Offset Correction: v_off_ini = 140.123 uV, v_off_fin = 136.878 uV 20:52:05 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 130.908 uVs 20:52:05 Integrated Voltage Measurement... 20:52:14 Voltage Offset Correction: v_off_ini = 136.191 uV, v_off_fin = 126.750 uV 20:52:14 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -139.388 uVs 20:52:14 Integrated Voltage Measurement... 20:52:23 Voltage Offset Correction: v_off_ini = 134.466 uV, v_off_fin = 127.638 uV 20:52:23 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 120.098 uVs 20:52:26 I2X Average and RMS of Measurements VT = -127.488 +- 2.310 uVs I2X = 11.729 +- 0.213 uTm^2 20:52:31 Wire Position: X0 = 1.0000 mm, Y0 = 0.0000 mm 20:52:33 Integrated Voltage Measurement... 20:52:42 Voltage Offset Correction: v_off_ini = 127.348 uV, v_off_fin = 117.205 uV 20:52:42 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -268.803 uVs 20:52:42 Integrated Voltage Measurement... 20:52:51 Voltage Offset Correction: v_off_ini = 113.547 uV, v_off_fin = 118.254 uV 20:52:51 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 252.195 uVs 20:52:51 Integrated Voltage Measurement... 20:53:00 Voltage Offset Correction: v_off_ini = 111.498 uV, v_off_fin = 108.078 uV 20:53:00 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -267.689 uVs 20:53:00 Integrated Voltage Measurement... 20:53:09 Voltage Offset Correction: v_off_ini = 119.247 uV, v_off_fin = 110.766 uV 20:53:09 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 277.315 uVs 20:53:09 Integrated Voltage Measurement... 20:53:18 Voltage Offset Correction: v_off_ini = 98.833 uV, v_off_fin = 96.532 uV 20:53:19 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -253.280 uVs 20:53:19 Integrated Voltage Measurement... 20:53:28 Voltage Offset Correction: v_off_ini = 114.784 uV, v_off_fin = 115.421 uV 20:53:28 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 243.702 uVs 20:53:30 I2Y Average and RMS of Measurements VT = -260.497 +- 9.802 uVs I2Y = -23.966 +- 0.902 uTm^2 20:53:35 Wire Position: X0 = 2.0000 mm, Y0 = 0.0000 mm 20:53:39 Integrated Voltage Measurement... 20:53:48 Voltage Offset Correction: v_off_ini = 123.962 uV, v_off_fin = 119.487 uV 20:53:48 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.4997 mm dXY = -0.4998 mm VT = -129.845 uVs 20:53:48 Integrated Voltage Measurement... 20:53:57 Voltage Offset Correction: v_off_ini = 120.639 uV, v_off_fin = 136.456 uV 20:53:57 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 125.329 uVs 20:53:57 Integrated Voltage Measurement... 20:54:06 Voltage Offset Correction: v_off_ini = 116.357 uV, v_off_fin = 104.169 uV 20:54:06 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -143.335 uVs 20:54:06 Integrated Voltage Measurement... 20:54:15 Voltage Offset Correction: v_off_ini = 113.109 uV, v_off_fin = 103.021 uV 20:54:15 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 140.987 uVs 20:54:15 Integrated Voltage Measurement... 20:54:24 Voltage Offset Correction: v_off_ini = 108.851 uV, v_off_fin = 105.695 uV 20:54:24 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -133.576 uVs 20:54:24 Integrated Voltage Measurement... 20:54:33 Voltage Offset Correction: v_off_ini = 111.335 uV, v_off_fin = 120.347 uV 20:54:33 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 141.495 uVs 20:54:37 I1X Average and RMS of Measurements VT = -135.761 +- 6.081 uVs I1X = 27.154 +- 1.214 uTm 20:54:40 Wire Position: X0 = 2.0000 mm, Y0 = 0.0000 mm 20:54:44 Integrated Voltage Measurement... 20:54:53 Voltage Offset Correction: v_off_ini = 114.238 uV, v_off_fin = 103.929 uV 20:54:53 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -192.796 uVs 20:54:53 Integrated Voltage Measurement... 20:55:02 Voltage Offset Correction: v_off_ini = 91.124 uV, v_off_fin = 108.868 uV 20:55:02 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 187.529 uVs 20:55:02 Integrated Voltage Measurement... 20:55:11 Voltage Offset Correction: v_off_ini = 96.886 uV, v_off_fin = 95.404 uV 20:55:11 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -172.139 uVs 20:55:11 Integrated Voltage Measurement... 20:55:20 Voltage Offset Correction: v_off_ini = 110.963 uV, v_off_fin = 102.179 uV 20:55:20 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 181.522 uVs 20:55:20 Integrated Voltage Measurement... 20:55:29 Voltage Offset Correction: v_off_ini = 92.830 uV, v_off_fin = 96.127 uV 20:55:29 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -180.323 uVs 20:55:29 Integrated Voltage Measurement... 20:55:38 Voltage Offset Correction: v_off_ini = 103.049 uV, v_off_fin = 106.555 uV 20:55:38 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 165.223 uVs 20:55:42 I1Y Average and RMS of Measurements VT = -179.922 +- 7.428 uVs I1Y = -35.984 +- 1.486 uTm 20:55:47 Wire Position: X0 = 2.0000 mm, Y0 = 0.0000 mm 20:55:50 Integrated Voltage Measurement... 20:55:59 Voltage Offset Correction: v_off_ini = 95.937 uV, v_off_fin = 97.502 uV 20:55:59 Integrated Voltage Measurement dXYus = -1.0001 mm, dXYds = 0.0000 mm dXY = -1.0001 mm VT = -128.557 uVs 20:55:59 Integrated Voltage Measurement... 20:56:08 Voltage Offset Correction: v_off_ini = 102.177 uV, v_off_fin = 97.883 uV 20:56:08 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 141.275 uVs 20:56:08 Integrated Voltage Measurement... 20:56:17 Voltage Offset Correction: v_off_ini = 85.935 uV, v_off_fin = 85.693 uV 20:56:17 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -144.511 uVs 20:56:17 Integrated Voltage Measurement... 20:56:26 Voltage Offset Correction: v_off_ini = 91.091 uV, v_off_fin = 100.571 uV 20:56:26 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 125.236 uVs 20:56:26 Integrated Voltage Measurement... 20:56:35 Voltage Offset Correction: v_off_ini = 82.468 uV, v_off_fin = 76.586 uV 20:56:35 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -134.674 uVs 20:56:35 Integrated Voltage Measurement... 20:56:44 Voltage Offset Correction: v_off_ini = 82.265 uV, v_off_fin = 90.725 uV 20:56:44 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 150.483 uVs 20:56:47 I2X Average and RMS of Measurements VT = -137.456 +- 3.622 uVs I2X = 12.646 +- 0.333 uTm^2 20:56:52 Wire Position: X0 = 2.0000 mm, Y0 = 0.0000 mm 20:56:54 Integrated Voltage Measurement... 20:57:03 Voltage Offset Correction: v_off_ini = 75.104 uV, v_off_fin = 76.752 uV 20:57:03 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -248.671 uVs 20:57:03 Integrated Voltage Measurement... 20:57:12 Voltage Offset Correction: v_off_ini = 80.779 uV, v_off_fin = 83.744 uV 20:57:12 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 261.946 uVs 20:57:12 Integrated Voltage Measurement... 20:57:21 Voltage Offset Correction: v_off_ini = 79.806 uV, v_off_fin = 83.506 uV 20:57:21 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -254.147 uVs 20:57:21 Integrated Voltage Measurement... 20:57:30 Voltage Offset Correction: v_off_ini = 94.842 uV, v_off_fin = 92.921 uV 20:57:30 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 227.957 uVs 20:57:31 Integrated Voltage Measurement... 20:57:39 Voltage Offset Correction: v_off_ini = 81.195 uV, v_off_fin = 78.246 uV 20:57:40 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -241.932 uVs 20:57:40 Integrated Voltage Measurement... 20:57:49 Voltage Offset Correction: v_off_ini = 85.047 uV, v_off_fin = 84.378 uV 20:57:49 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 252.951 uVs 20:57:51 I2Y Average and RMS of Measurements VT = -247.934 +- 5.831 uVs I2Y = -22.810 +- 0.536 uTm^2 20:57:59 Wire Position: X0 = 0.0000 mm, Y0 = -1.0000 mm 20:58:03 Integrated Voltage Measurement... 20:58:12 Voltage Offset Correction: v_off_ini = 82.681 uV, v_off_fin = 65.688 uV 20:58:12 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -110.281 uVs 20:58:12 Integrated Voltage Measurement... 20:58:21 Voltage Offset Correction: v_off_ini = 69.416 uV, v_off_fin = 64.033 uV 20:58:21 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 99.037 uVs 20:58:21 Integrated Voltage Measurement... 20:58:30 Voltage Offset Correction: v_off_ini = 71.974 uV, v_off_fin = 63.719 uV 20:58:30 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -118.187 uVs 20:58:30 Integrated Voltage Measurement... 20:58:39 Voltage Offset Correction: v_off_ini = 58.939 uV, v_off_fin = 66.828 uV 20:58:39 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 124.153 uVs 20:58:39 Integrated Voltage Measurement... 20:58:48 Voltage Offset Correction: v_off_ini = 78.040 uV, v_off_fin = 75.769 uV 20:58:48 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -106.934 uVs 20:58:48 Integrated Voltage Measurement... 20:58:57 Voltage Offset Correction: v_off_ini = 71.122 uV, v_off_fin = 78.616 uV 20:58:57 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 137.465 uVs 20:59:01 I1X Average and RMS of Measurements VT = -116.009 +- 8.037 uVs I1X = 23.203 +- 1.608 uTm 20:59:04 Wire Position: X0 = 0.0000 mm, Y0 = -1.0000 mm 20:59:08 Integrated Voltage Measurement... 20:59:17 Voltage Offset Correction: v_off_ini = 86.692 uV, v_off_fin = 77.866 uV 20:59:17 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -184.143 uVs 20:59:17 Integrated Voltage Measurement... 20:59:26 Voltage Offset Correction: v_off_ini = 91.123 uV, v_off_fin = 100.761 uV 20:59:26 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 176.599 uVs 20:59:26 Integrated Voltage Measurement... 20:59:35 Voltage Offset Correction: v_off_ini = 94.050 uV, v_off_fin = 84.433 uV 20:59:35 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -183.487 uVs 20:59:35 Integrated Voltage Measurement... 20:59:44 Voltage Offset Correction: v_off_ini = 78.188 uV, v_off_fin = 93.959 uV 20:59:44 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 169.668 uVs 20:59:44 Integrated Voltage Measurement... 20:59:53 Voltage Offset Correction: v_off_ini = 84.108 uV, v_off_fin = 76.537 uV 20:59:53 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -162.197 uVs 20:59:53 Integrated Voltage Measurement... 21:00:02 Voltage Offset Correction: v_off_ini = 92.429 uV, v_off_fin = 94.363 uV 21:00:02 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 172.827 uVs 21:00:06 I1Y Average and RMS of Measurements VT = -174.820 +- 5.395 uVs I1Y = -34.964 +- 1.079 uTm 21:00:09 Wire Position: X0 = 0.0000 mm, Y0 = -1.0000 mm 21:00:11 Integrated Voltage Measurement... 21:00:20 Voltage Offset Correction: v_off_ini = 63.918 uV, v_off_fin = 68.544 uV 21:00:20 Integrated Voltage Measurement dXYus = -1.0001 mm, dXYds = 0.0000 mm dXY = -1.0001 mm VT = -112.573 uVs 21:00:20 Integrated Voltage Measurement... 21:00:29 Voltage Offset Correction: v_off_ini = 79.985 uV, v_off_fin = 81.597 uV 21:00:29 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 117.959 uVs 21:00:30 Integrated Voltage Measurement... 21:00:38 Voltage Offset Correction: v_off_ini = 86.309 uV, v_off_fin = 64.261 uV 21:00:39 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -127.257 uVs 21:00:39 Integrated Voltage Measurement... 21:00:48 Voltage Offset Correction: v_off_ini = 83.237 uV, v_off_fin = 83.160 uV 21:00:48 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 109.037 uVs 21:00:48 Integrated Voltage Measurement... 21:00:57 Voltage Offset Correction: v_off_ini = 74.172 uV, v_off_fin = 69.891 uV 21:00:57 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -145.634 uVs 21:00:57 Integrated Voltage Measurement... 21:01:06 Voltage Offset Correction: v_off_ini = 76.810 uV, v_off_fin = 85.415 uV 21:01:06 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 96.221 uVs 21:01:08 I2X Average and RMS of Measurements VT = -118.114 +- 2.311 uVs I2X = 10.866 +- 0.213 uTm^2 21:01:11 Wire Position: X0 = 0.0000 mm, Y0 = -1.0000 mm 21:01:13 Integrated Voltage Measurement... 21:01:22 Voltage Offset Correction: v_off_ini = 66.602 uV, v_off_fin = 61.700 uV 21:01:22 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -254.797 uVs 21:01:22 Integrated Voltage Measurement... 21:01:31 Voltage Offset Correction: v_off_ini = 73.685 uV, v_off_fin = 87.084 uV 21:01:31 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 262.280 uVs 21:01:31 Integrated Voltage Measurement... 21:01:40 Voltage Offset Correction: v_off_ini = 88.253 uV, v_off_fin = 96.040 uV 21:01:40 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -243.055 uVs 21:01:40 Integrated Voltage Measurement... 21:01:49 Voltage Offset Correction: v_off_ini = 86.252 uV, v_off_fin = 93.307 uV 21:01:49 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 248.648 uVs 21:01:49 Integrated Voltage Measurement... 21:01:58 Voltage Offset Correction: v_off_ini = 87.849 uV, v_off_fin = 93.650 uV 21:01:58 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -260.966 uVs 21:01:58 Integrated Voltage Measurement... 21:02:07 Voltage Offset Correction: v_off_ini = 95.541 uV, v_off_fin = 99.880 uV 21:02:07 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 258.489 uVs 21:02:10 I2Y Average and RMS of Measurements VT = -254.706 +- 6.280 uVs I2Y = -23.433 +- 0.578 uTm^2 21:02:14 Wire Position: X0 = 0.0000 mm, Y0 = -0.5000 mm 21:02:18 Integrated Voltage Measurement... 21:02:27 Voltage Offset Correction: v_off_ini = 106.202 uV, v_off_fin = 100.820 uV 21:02:27 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -120.981 uVs 21:02:27 Integrated Voltage Measurement... 21:02:36 Voltage Offset Correction: v_off_ini = 104.086 uV, v_off_fin = 112.651 uV 21:02:36 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 113.600 uVs 21:02:36 Integrated Voltage Measurement... 21:02:45 Voltage Offset Correction: v_off_ini = 109.932 uV, v_off_fin = 97.125 uV 21:02:45 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -116.900 uVs 21:02:45 Integrated Voltage Measurement... 21:02:54 Voltage Offset Correction: v_off_ini = 100.189 uV, v_off_fin = 99.600 uV 21:02:54 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = 114.116 uVs 21:02:54 Integrated Voltage Measurement... 21:03:03 Voltage Offset Correction: v_off_ini = 98.052 uV, v_off_fin = 99.761 uV 21:03:03 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -131.299 uVs 21:03:03 Integrated Voltage Measurement... 21:03:12 Voltage Offset Correction: v_off_ini = 117.892 uV, v_off_fin = 110.519 uV 21:03:12 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 119.253 uVs 21:03:16 I1X Average and RMS of Measurements VT = -119.358 +- 4.247 uVs I1X = 23.872 +- 0.849 uTm 21:03:19 Wire Position: X0 = 0.0000 mm, Y0 = -0.5000 mm 21:03:23 Integrated Voltage Measurement... 21:03:32 Voltage Offset Correction: v_off_ini = 101.317 uV, v_off_fin = 103.622 uV 21:03:32 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -178.788 uVs 21:03:32 Integrated Voltage Measurement... 21:03:41 Voltage Offset Correction: v_off_ini = 109.343 uV, v_off_fin = 115.202 uV 21:03:41 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 176.640 uVs 21:03:41 Integrated Voltage Measurement... 21:03:50 Voltage Offset Correction: v_off_ini = 113.503 uV, v_off_fin = 106.150 uV 21:03:50 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -175.345 uVs 21:03:50 Integrated Voltage Measurement... 21:03:59 Voltage Offset Correction: v_off_ini = 116.218 uV, v_off_fin = 119.247 uV 21:03:59 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 168.320 uVs 21:03:59 Integrated Voltage Measurement... 21:04:08 Voltage Offset Correction: v_off_ini = 116.020 uV, v_off_fin = 104.200 uV 21:04:08 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -185.317 uVs 21:04:08 Integrated Voltage Measurement... 21:04:17 Voltage Offset Correction: v_off_ini = 94.050 uV, v_off_fin = 108.426 uV 21:04:17 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 186.495 uVs 21:04:21 I1Y Average and RMS of Measurements VT = -178.484 +- 5.771 uVs I1Y = -35.697 +- 1.154 uTm 21:04:24 Wire Position: X0 = 0.0000 mm, Y0 = -0.5000 mm 21:04:27 Integrated Voltage Measurement... 21:04:36 Voltage Offset Correction: v_off_ini = 97.865 uV, v_off_fin = 93.931 uV 21:04:36 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -129.403 uVs 21:04:36 Integrated Voltage Measurement... 21:04:45 Voltage Offset Correction: v_off_ini = 94.821 uV, v_off_fin = 115.409 uV 21:04:45 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 136.069 uVs 21:04:45 Integrated Voltage Measurement... 21:04:54 Voltage Offset Correction: v_off_ini = 114.859 uV, v_off_fin = 112.602 uV 21:04:54 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -134.686 uVs 21:04:54 Integrated Voltage Measurement... 21:05:03 Voltage Offset Correction: v_off_ini = 111.806 uV, v_off_fin = 116.175 uV 21:05:03 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 134.398 uVs 21:05:03 Integrated Voltage Measurement... 21:05:12 Voltage Offset Correction: v_off_ini = 106.056 uV, v_off_fin = 94.314 uV 21:05:12 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -119.950 uVs 21:05:12 Integrated Voltage Measurement... 21:05:21 Voltage Offset Correction: v_off_ini = 110.954 uV, v_off_fin = 121.021 uV 21:05:21 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 136.363 uVs 21:05:23 I2X Average and RMS of Measurements VT = -131.812 +- 2.688 uVs I2X = 12.127 +- 0.247 uTm^2 21:05:26 Wire Position: X0 = 0.0000 mm, Y0 = -0.5000 mm 21:05:28 Integrated Voltage Measurement... 21:05:37 Voltage Offset Correction: v_off_ini = 94.195 uV, v_off_fin = 97.795 uV 21:05:37 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -262.845 uVs 21:05:37 Integrated Voltage Measurement... 21:05:46 Voltage Offset Correction: v_off_ini = 94.695 uV, v_off_fin = 96.494 uV 21:05:46 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = 248.569 uVs 21:05:46 Integrated Voltage Measurement... 21:05:55 Voltage Offset Correction: v_off_ini = 108.559 uV, v_off_fin = 84.369 uV 21:05:55 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -261.865 uVs 21:05:55 Integrated Voltage Measurement... 21:06:04 Voltage Offset Correction: v_off_ini = 92.042 uV, v_off_fin = 89.811 uV 21:06:05 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 256.139 uVs 21:06:05 Integrated Voltage Measurement... 21:06:14 Voltage Offset Correction: v_off_ini = 79.087 uV, v_off_fin = 85.611 uV 21:06:14 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -258.609 uVs 21:06:14 Integrated Voltage Measurement... 21:06:23 Voltage Offset Correction: v_off_ini = 90.355 uV, v_off_fin = 94.529 uV 21:06:23 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 257.914 uVs 21:06:25 I2Y Average and RMS of Measurements VT = -257.657 +- 1.411 uVs I2Y = -23.705 +- 0.129 uTm^2 21:06:29 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 21:06:33 Integrated Voltage Measurement... 21:06:42 Voltage Offset Correction: v_off_ini = 88.367 uV, v_off_fin = 90.214 uV 21:06:42 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.4998 mm dXY = -0.4999 mm VT = -125.402 uVs 21:06:42 Integrated Voltage Measurement... 21:06:51 Voltage Offset Correction: v_off_ini = 93.199 uV, v_off_fin = 86.600 uV 21:06:51 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 135.372 uVs 21:06:51 Integrated Voltage Measurement... 21:07:00 Voltage Offset Correction: v_off_ini = 84.207 uV, v_off_fin = 79.932 uV 21:07:00 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.4999 mm dXY = -0.4999 mm VT = -141.599 uVs 21:07:00 Integrated Voltage Measurement... 21:07:09 Voltage Offset Correction: v_off_ini = 87.687 uV, v_off_fin = 84.011 uV 21:07:09 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 131.672 uVs 21:07:09 Integrated Voltage Measurement... 21:07:18 Voltage Offset Correction: v_off_ini = 85.007 uV, v_off_fin = 88.254 uV 21:07:18 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -120.794 uVs 21:07:19 Integrated Voltage Measurement... 21:07:27 Voltage Offset Correction: v_off_ini = 79.175 uV, v_off_fin = 87.000 uV 21:07:27 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 146.255 uVs 21:07:32 I1X Average and RMS of Measurements VT = -133.516 +- 2.551 uVs I1X = 26.706 +- 0.509 uTm 21:07:34 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 21:07:39 Integrated Voltage Measurement... 21:07:47 Voltage Offset Correction: v_off_ini = 89.905 uV, v_off_fin = 91.915 uV 21:07:47 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -174.018 uVs 21:07:48 Integrated Voltage Measurement... 21:07:56 Voltage Offset Correction: v_off_ini = 106.406 uV, v_off_fin = 124.063 uV 21:07:56 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 179.763 uVs 21:07:57 Integrated Voltage Measurement... 21:08:05 Voltage Offset Correction: v_off_ini = 101.424 uV, v_off_fin = 99.140 uV 21:08:05 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -176.457 uVs 21:08:06 Integrated Voltage Measurement... 21:08:14 Voltage Offset Correction: v_off_ini = 96.264 uV, v_off_fin = 115.611 uV 21:08:14 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 165.726 uVs 21:08:15 Integrated Voltage Measurement... 21:08:23 Voltage Offset Correction: v_off_ini = 97.839 uV, v_off_fin = 92.114 uV 21:08:24 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -177.561 uVs 21:08:24 Integrated Voltage Measurement... 21:08:32 Voltage Offset Correction: v_off_ini = 107.062 uV, v_off_fin = 113.297 uV 21:08:33 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 182.137 uVs 21:08:37 I1Y Average and RMS of Measurements VT = -175.944 +- 3.638 uVs I1Y = -35.189 +- 0.728 uTm 21:08:39 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 21:08:42 Integrated Voltage Measurement... 21:08:51 Voltage Offset Correction: v_off_ini = 104.963 uV, v_off_fin = 91.674 uV 21:08:51 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -125.610 uVs 21:08:51 Integrated Voltage Measurement... 21:09:00 Voltage Offset Correction: v_off_ini = 109.610 uV, v_off_fin = 113.355 uV 21:09:00 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 129.343 uVs 21:09:00 Integrated Voltage Measurement... 21:09:09 Voltage Offset Correction: v_off_ini = 105.767 uV, v_off_fin = 87.627 uV 21:09:09 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -147.336 uVs 21:09:09 Integrated Voltage Measurement... 21:09:18 Voltage Offset Correction: v_off_ini = 104.491 uV, v_off_fin = 89.734 uV 21:09:18 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 116.496 uVs 21:09:18 Integrated Voltage Measurement... 21:09:27 Voltage Offset Correction: v_off_ini = 72.999 uV, v_off_fin = 74.075 uV 21:09:27 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -131.301 uVs 21:09:27 Integrated Voltage Measurement... 21:09:36 Voltage Offset Correction: v_off_ini = 83.217 uV, v_off_fin = 89.987 uV 21:09:36 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 118.944 uVs 21:09:38 I2X Average and RMS of Measurements VT = -128.172 +- 2.817 uVs I2X = 11.792 +- 0.259 uTm^2 21:09:41 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 21:09:44 Integrated Voltage Measurement... 21:09:53 Voltage Offset Correction: v_off_ini = 85.927 uV, v_off_fin = 90.594 uV 21:09:53 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -249.812 uVs 21:09:53 Integrated Voltage Measurement... 21:10:02 Voltage Offset Correction: v_off_ini = 97.034 uV, v_off_fin = 89.478 uV 21:10:02 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 264.338 uVs 21:10:02 Integrated Voltage Measurement... 21:10:11 Voltage Offset Correction: v_off_ini = 89.044 uV, v_off_fin = 99.759 uV 21:10:11 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -275.814 uVs 21:10:11 Integrated Voltage Measurement... 21:10:20 Voltage Offset Correction: v_off_ini = 96.152 uV, v_off_fin = 83.584 uV 21:10:20 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 262.175 uVs 21:10:20 Integrated Voltage Measurement... 21:10:29 Voltage Offset Correction: v_off_ini = 107.642 uV, v_off_fin = 93.316 uV 21:10:29 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -256.252 uVs 21:10:29 Integrated Voltage Measurement... 21:10:38 Voltage Offset Correction: v_off_ini = 86.258 uV, v_off_fin = 102.551 uV 21:10:38 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 268.099 uVs 21:10:40 I2Y Average and RMS of Measurements VT = -262.748 +- 4.883 uVs I2Y = -24.173 +- 0.449 uTm^2 21:10:44 Wire Position: X0 = 0.0000 mm, Y0 = 0.5000 mm 21:10:49 Integrated Voltage Measurement... 21:10:58 Voltage Offset Correction: v_off_ini = 83.514 uV, v_off_fin = 85.073 uV 21:10:58 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -138.658 uVs 21:10:58 Integrated Voltage Measurement... 21:11:07 Voltage Offset Correction: v_off_ini = 72.389 uV, v_off_fin = 77.356 uV 21:11:07 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 140.787 uVs 21:11:07 Integrated Voltage Measurement... 21:11:16 Voltage Offset Correction: v_off_ini = 80.915 uV, v_off_fin = 69.401 uV 21:11:16 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -124.973 uVs 21:11:16 Integrated Voltage Measurement... 21:11:25 Voltage Offset Correction: v_off_ini = 66.482 uV, v_off_fin = 77.009 uV 21:11:25 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 118.338 uVs 21:11:25 Integrated Voltage Measurement... 21:11:34 Voltage Offset Correction: v_off_ini = 76.778 uV, v_off_fin = 70.243 uV 21:11:34 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -120.151 uVs 21:11:34 Integrated Voltage Measurement... 21:11:43 Voltage Offset Correction: v_off_ini = 79.926 uV, v_off_fin = 71.383 uV 21:11:43 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 124.086 uVs 21:11:47 I1X Average and RMS of Measurements VT = -127.832 +- 8.410 uVs I1X = 25.567 +- 1.682 uTm 21:11:49 Wire Position: X0 = 0.0000 mm, Y0 = 0.5000 mm 21:11:54 Integrated Voltage Measurement... 21:12:03 Voltage Offset Correction: v_off_ini = 77.840 uV, v_off_fin = 70.819 uV 21:12:03 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -176.268 uVs 21:12:03 Integrated Voltage Measurement... 21:12:12 Voltage Offset Correction: v_off_ini = 77.380 uV, v_off_fin = 85.248 uV 21:12:12 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 177.984 uVs 21:12:12 Integrated Voltage Measurement... 21:12:21 Voltage Offset Correction: v_off_ini = 84.368 uV, v_off_fin = 74.858 uV 21:12:21 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -175.778 uVs 21:12:21 Integrated Voltage Measurement... 21:12:30 Voltage Offset Correction: v_off_ini = 87.843 uV, v_off_fin = 97.144 uV 21:12:30 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 185.240 uVs 21:12:30 Integrated Voltage Measurement... 21:12:39 Voltage Offset Correction: v_off_ini = 65.421 uV, v_off_fin = 61.751 uV 21:12:39 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -164.606 uVs 21:12:39 Integrated Voltage Measurement... 21:12:48 Voltage Offset Correction: v_off_ini = 64.533 uV, v_off_fin = 88.302 uV 21:12:48 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 171.848 uVs 21:12:52 I1Y Average and RMS of Measurements VT = -175.287 +- 5.180 uVs I1Y = -35.057 +- 1.036 uTm 21:12:54 Wire Position: X0 = 0.0000 mm, Y0 = 0.5000 mm 21:12:57 Integrated Voltage Measurement... 21:13:06 Voltage Offset Correction: v_off_ini = 64.802 uV, v_off_fin = 56.292 uV 21:13:06 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -135.246 uVs 21:13:06 Integrated Voltage Measurement... 21:13:15 Voltage Offset Correction: v_off_ini = 67.832 uV, v_off_fin = 75.450 uV 21:13:15 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 129.488 uVs 21:13:15 Integrated Voltage Measurement... 21:13:24 Voltage Offset Correction: v_off_ini = 64.301 uV, v_off_fin = 60.983 uV 21:13:24 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -126.123 uVs 21:13:24 Integrated Voltage Measurement... 21:13:33 Voltage Offset Correction: v_off_ini = 65.264 uV, v_off_fin = 59.664 uV 21:13:33 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 130.753 uVs 21:13:33 Integrated Voltage Measurement... 21:13:42 Voltage Offset Correction: v_off_ini = 52.776 uV, v_off_fin = 51.776 uV 21:13:42 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -132.959 uVs 21:13:42 Integrated Voltage Measurement... 21:13:51 Voltage Offset Correction: v_off_ini = 62.741 uV, v_off_fin = 70.553 uV 21:13:51 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 135.597 uVs 21:13:54 I2X Average and RMS of Measurements VT = -131.694 +- 2.431 uVs I2X = 12.116 +- 0.224 uTm^2 21:13:56 Wire Position: X0 = 0.0000 mm, Y0 = 0.5000 mm 21:13:59 Integrated Voltage Measurement... 21:14:08 Voltage Offset Correction: v_off_ini = 49.591 uV, v_off_fin = 57.744 uV 21:14:08 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -268.256 uVs 21:14:08 Integrated Voltage Measurement... 21:14:17 Voltage Offset Correction: v_off_ini = 59.293 uV, v_off_fin = 58.219 uV 21:14:17 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 279.495 uVs 21:14:17 Integrated Voltage Measurement... 21:14:26 Voltage Offset Correction: v_off_ini = 55.561 uV, v_off_fin = 44.001 uV 21:14:26 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -262.626 uVs 21:14:26 Integrated Voltage Measurement... 21:14:35 Voltage Offset Correction: v_off_ini = 62.114 uV, v_off_fin = 33.719 uV 21:14:35 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 259.394 uVs 21:14:35 Integrated Voltage Measurement... 21:14:44 Voltage Offset Correction: v_off_ini = 39.736 uV, v_off_fin = 34.078 uV 21:14:44 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -249.360 uVs 21:14:44 Integrated Voltage Measurement... 21:14:53 Voltage Offset Correction: v_off_ini = 27.317 uV, v_off_fin = 36.876 uV 21:14:53 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = 268.945 uVs 21:14:55 I2Y Average and RMS of Measurements VT = -264.679 +- 6.547 uVs I2Y = -24.351 +- 0.602 uTm^2 21:14:59 Wire Position: X0 = 0.0000 mm, Y0 = 1.0000 mm 21:15:04 Integrated Voltage Measurement... 21:15:13 Voltage Offset Correction: v_off_ini = 36.644 uV, v_off_fin = 31.459 uV 21:15:13 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -131.459 uVs 21:15:13 Integrated Voltage Measurement... 21:15:22 Voltage Offset Correction: v_off_ini = 37.891 uV, v_off_fin = 35.259 uV 21:15:22 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 111.196 uVs 21:15:22 Integrated Voltage Measurement... 21:15:31 Voltage Offset Correction: v_off_ini = 31.333 uV, v_off_fin = 32.304 uV 21:15:31 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -117.668 uVs 21:15:31 Integrated Voltage Measurement... 21:15:40 Voltage Offset Correction: v_off_ini = 40.888 uV, v_off_fin = 38.219 uV 21:15:40 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 106.840 uVs 21:15:40 Integrated Voltage Measurement... 21:15:49 Voltage Offset Correction: v_off_ini = 36.438 uV, v_off_fin = 31.136 uV 21:15:49 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -105.169 uVs 21:15:49 Integrated Voltage Measurement... 21:15:58 Voltage Offset Correction: v_off_ini = 44.658 uV, v_off_fin = 46.602 uV 21:15:58 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 124.188 uVs 21:16:02 I1X Average and RMS of Measurements VT = -116.087 +- 3.836 uVs I1X = 23.217 +- 0.767 uTm 21:16:04 Wire Position: X0 = 0.0000 mm, Y0 = 1.0000 mm 21:16:09 Integrated Voltage Measurement... 21:16:18 Voltage Offset Correction: v_off_ini = 37.556 uV, v_off_fin = 29.902 uV 21:16:18 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -177.256 uVs 21:16:18 Integrated Voltage Measurement... 21:16:27 Voltage Offset Correction: v_off_ini = 33.126 uV, v_off_fin = 37.911 uV 21:16:27 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 178.691 uVs 21:16:27 Integrated Voltage Measurement... 21:16:36 Voltage Offset Correction: v_off_ini = 22.918 uV, v_off_fin = 3.353 uV 21:16:36 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -166.427 uVs 21:16:36 Integrated Voltage Measurement... 21:16:45 Voltage Offset Correction: v_off_ini = 25.000 uV, v_off_fin = 31.797 uV 21:16:45 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 183.146 uVs 21:16:45 Integrated Voltage Measurement... 21:16:54 Voltage Offset Correction: v_off_ini = 21.592 uV, v_off_fin = 15.531 uV 21:16:54 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -193.085 uVs 21:16:54 Integrated Voltage Measurement... 21:17:03 Voltage Offset Correction: v_off_ini = 21.782 uV, v_off_fin = 32.473 uV 21:17:03 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 171.503 uVs 21:17:07 I1Y Average and RMS of Measurements VT = -178.351 +- 3.077 uVs I1Y = -35.668 +- 0.614 uTm 21:17:09 Wire Position: X0 = 0.0000 mm, Y0 = 1.0000 mm 21:17:12 Integrated Voltage Measurement... 21:17:21 Voltage Offset Correction: v_off_ini = 26.557 uV, v_off_fin = 9.282 uV 21:17:21 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -117.631 uVs 21:17:21 Integrated Voltage Measurement... 21:17:30 Voltage Offset Correction: v_off_ini = 15.512 uV, v_off_fin = 25.308 uV 21:17:30 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 119.624 uVs 21:17:30 Integrated Voltage Measurement... 21:17:39 Voltage Offset Correction: v_off_ini = 26.454 uV, v_off_fin = 6.015 uV 21:17:39 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -132.131 uVs 21:17:39 Integrated Voltage Measurement... 21:17:48 Voltage Offset Correction: v_off_ini = 18.518 uV, v_off_fin = 27.815 uV 21:17:48 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 138.707 uVs 21:17:48 Integrated Voltage Measurement... 21:17:57 Voltage Offset Correction: v_off_ini = 17.860 uV, v_off_fin = 8.584 uV 21:17:57 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -143.882 uVs 21:17:57 Integrated Voltage Measurement... 21:18:06 Voltage Offset Correction: v_off_ini = 13.936 uV, v_off_fin = 26.656 uV 21:18:06 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 133.247 uVs 21:18:09 I2X Average and RMS of Measurements VT = -130.871 +- 8.752 uVs I2X = 12.041 +- 0.805 uTm^2 21:18:11 Wire Position: X0 = 0.0000 mm, Y0 = 1.0000 mm 21:18:14 Integrated Voltage Measurement... 21:18:23 Voltage Offset Correction: v_off_ini = 32.635 uV, v_off_fin = 22.630 uV 21:18:23 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -265.433 uVs 21:18:23 Integrated Voltage Measurement... 21:18:32 Voltage Offset Correction: v_off_ini = 11.850 uV, v_off_fin = 17.650 uV 21:18:32 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 270.940 uVs 21:18:32 Integrated Voltage Measurement... 21:18:41 Voltage Offset Correction: v_off_ini = 15.089 uV, v_off_fin = 26.530 uV 21:18:41 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -243.549 uVs 21:18:41 Integrated Voltage Measurement... 21:18:50 Voltage Offset Correction: v_off_ini = 21.915 uV, v_off_fin = 16.563 uV 21:18:50 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 265.379 uVs 21:18:50 Integrated Voltage Measurement... 21:18:59 Voltage Offset Correction: v_off_ini = 33.335 uV, v_off_fin = 28.985 uV 21:18:59 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -265.144 uVs 21:18:59 Integrated Voltage Measurement... 21:19:08 Voltage Offset Correction: v_off_ini = 36.935 uV, v_off_fin = 30.849 uV 21:19:08 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 255.893 uVs 21:19:10 I2Y Average and RMS of Measurements VT = -261.056 +- 5.615 uVs I2Y = -24.017 +- 0.517 uTm^2 21:19:42 Wire Position: X0 = -2.0000 mm, Y0 = 0.0000 mm 21:19:46 Integrated Voltage Measurement... 21:19:55 Voltage Offset Correction: v_off_ini = 21.629 uV, v_off_fin = 16.400 uV 21:19:55 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.4998 mm dXY = -0.4999 mm VT = -112.096 uVs 21:19:55 Integrated Voltage Measurement... 21:20:04 Voltage Offset Correction: v_off_ini = 28.233 uV, v_off_fin = 23.887 uV 21:20:04 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = 76.101 uVs 21:20:04 Integrated Voltage Measurement... 21:20:13 Voltage Offset Correction: v_off_ini = 15.768 uV, v_off_fin = 15.741 uV 21:20:13 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -76.587 uVs 21:20:13 Integrated Voltage Measurement... 21:20:22 Voltage Offset Correction: v_off_ini = 24.497 uV, v_off_fin = 35.195 uV 21:20:22 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = 98.938 uVs 21:20:22 Integrated Voltage Measurement... 21:20:31 Voltage Offset Correction: v_off_ini = 22.485 uV, v_off_fin = 20.172 uV 21:20:31 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -102.379 uVs 21:20:31 Integrated Voltage Measurement... 21:20:40 Voltage Offset Correction: v_off_ini = 20.618 uV, v_off_fin = 24.182 uV 21:20:40 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 101.796 uVs 21:20:44 I1X Average and RMS of Measurements VT = -94.649 +- 5.861 uVs I1X = 18.932 +- 1.172 uTm 21:20:47 Wire Position: X0 = -2.0000 mm, Y0 = 0.0000 mm 21:20:51 Integrated Voltage Measurement... 21:21:00 Voltage Offset Correction: v_off_ini = 14.299 uV, v_off_fin = 11.305 uV 21:21:00 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -205.413 uVs 21:21:00 Integrated Voltage Measurement... 21:21:09 Voltage Offset Correction: v_off_ini = 25.682 uV, v_off_fin = 38.553 uV 21:21:09 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 209.085 uVs 21:21:09 Integrated Voltage Measurement... 21:21:18 Voltage Offset Correction: v_off_ini = 21.407 uV, v_off_fin = 18.729 uV 21:21:18 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -211.156 uVs 21:21:18 Integrated Voltage Measurement... 21:21:27 Voltage Offset Correction: v_off_ini = 15.589 uV, v_off_fin = 34.466 uV 21:21:27 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 216.200 uVs 21:21:27 Integrated Voltage Measurement... 21:21:36 Voltage Offset Correction: v_off_ini = 24.921 uV, v_off_fin = 10.142 uV 21:21:36 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -200.566 uVs 21:21:36 Integrated Voltage Measurement... 21:21:45 Voltage Offset Correction: v_off_ini = 15.661 uV, v_off_fin = 30.070 uV 21:21:45 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 210.160 uVs 21:21:49 I1Y Average and RMS of Measurements VT = -208.763 +- 3.559 uVs I1Y = -41.753 +- 0.712 uTm 21:21:55 Wire Position: X0 = -2.0000 mm, Y0 = 0.0000 mm 21:21:57 Integrated Voltage Measurement... 21:22:06 Voltage Offset Correction: v_off_ini = 37.905 uV, v_off_fin = 34.789 uV 21:22:06 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -102.076 uVs 21:22:06 Integrated Voltage Measurement... 21:22:15 Voltage Offset Correction: v_off_ini = 35.510 uV, v_off_fin = 42.753 uV 21:22:15 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 90.807 uVs 21:22:15 Integrated Voltage Measurement... 21:22:24 Voltage Offset Correction: v_off_ini = 16.018 uV, v_off_fin = 17.069 uV 21:22:24 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -80.601 uVs 21:22:24 Integrated Voltage Measurement... 21:22:33 Voltage Offset Correction: v_off_ini = 30.066 uV, v_off_fin = 30.036 uV 21:22:33 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 78.913 uVs 21:22:33 Integrated Voltage Measurement... 21:22:42 Voltage Offset Correction: v_off_ini = 25.335 uV, v_off_fin = 27.237 uV 21:22:42 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -98.482 uVs 21:22:42 Integrated Voltage Measurement... 21:22:51 Voltage Offset Correction: v_off_ini = 29.077 uV, v_off_fin = 41.027 uV 21:22:51 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 99.243 uVs 21:22:54 I2X Average and RMS of Measurements VT = -91.687 +- 8.493 uVs I2X = 8.435 +- 0.781 uTm^2 21:22:59 Wire Position: X0 = -2.0000 mm, Y0 = 0.0000 mm 21:23:01 Integrated Voltage Measurement... 21:23:10 Voltage Offset Correction: v_off_ini = 21.914 uV, v_off_fin = 15.829 uV 21:23:11 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -242.131 uVs 21:23:11 Integrated Voltage Measurement... 21:23:20 Voltage Offset Correction: v_off_ini = 17.567 uV, v_off_fin = 18.988 uV 21:23:20 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 245.204 uVs 21:23:20 Integrated Voltage Measurement... 21:23:29 Voltage Offset Correction: v_off_ini = 23.256 uV, v_off_fin = 26.620 uV 21:23:29 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -237.155 uVs 21:23:29 Integrated Voltage Measurement... 21:23:38 Voltage Offset Correction: v_off_ini = 23.863 uV, v_off_fin = 34.298 uV 21:23:38 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 240.098 uVs 21:23:38 Integrated Voltage Measurement... 21:23:47 Voltage Offset Correction: v_off_ini = 32.374 uV, v_off_fin = 29.801 uV 21:23:47 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -253.237 uVs 21:23:47 Integrated Voltage Measurement... 21:23:56 Voltage Offset Correction: v_off_ini = 47.867 uV, v_off_fin = 28.750 uV 21:23:56 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 216.604 uVs 21:23:58 I2Y Average and RMS of Measurements VT = -239.071 +- 3.585 uVs I2Y = -21.995 +- 0.330 uTm^2 21:24:03 Wire Position: X0 = -1.0000 mm, Y0 = 0.0000 mm 21:24:07 Integrated Voltage Measurement... 21:24:16 Voltage Offset Correction: v_off_ini = 36.966 uV, v_off_fin = 30.155 uV 21:24:16 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.4999 mm dXY = -0.4999 mm VT = -93.054 uVs 21:24:16 Integrated Voltage Measurement... 21:24:25 Voltage Offset Correction: v_off_ini = 32.483 uV, v_off_fin = 39.586 uV 21:24:25 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 82.156 uVs 21:24:25 Integrated Voltage Measurement... 21:24:34 Voltage Offset Correction: v_off_ini = 32.430 uV, v_off_fin = 21.316 uV 21:24:34 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -85.741 uVs 21:24:34 Integrated Voltage Measurement... 21:24:43 Voltage Offset Correction: v_off_ini = 38.706 uV, v_off_fin = 33.962 uV 21:24:43 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 88.858 uVs 21:24:43 Integrated Voltage Measurement... 21:24:52 Voltage Offset Correction: v_off_ini = 25.701 uV, v_off_fin = 10.164 uV 21:24:52 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -116.137 uVs 21:24:52 Integrated Voltage Measurement... 21:25:01 Voltage Offset Correction: v_off_ini = 21.285 uV, v_off_fin = 27.829 uV 21:25:01 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 106.414 uVs 21:25:05 I1X Average and RMS of Measurements VT = -95.393 +- 11.231 uVs I1X = 19.079 +- 2.246 uTm 21:25:08 Wire Position: X0 = -1.0000 mm, Y0 = 0.0000 mm 21:25:12 Integrated Voltage Measurement... 21:25:21 Voltage Offset Correction: v_off_ini = 32.524 uV, v_off_fin = 19.979 uV 21:25:21 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -225.102 uVs 21:25:21 Integrated Voltage Measurement... 21:25:30 Voltage Offset Correction: v_off_ini = 29.043 uV, v_off_fin = 47.122 uV 21:25:30 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 207.447 uVs 21:25:30 Integrated Voltage Measurement... 21:25:39 Voltage Offset Correction: v_off_ini = 38.543 uV, v_off_fin = 19.769 uV 21:25:39 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -219.007 uVs 21:25:39 Integrated Voltage Measurement... 21:25:48 Voltage Offset Correction: v_off_ini = 24.058 uV, v_off_fin = 30.769 uV 21:25:48 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 211.264 uVs 21:25:48 Integrated Voltage Measurement... 21:25:57 Voltage Offset Correction: v_off_ini = 34.699 uV, v_off_fin = 10.215 uV 21:25:57 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -222.450 uVs 21:25:57 Integrated Voltage Measurement... 21:26:06 Voltage Offset Correction: v_off_ini = 31.434 uV, v_off_fin = 31.767 uV 21:26:06 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 205.980 uVs 21:26:10 I1Y Average and RMS of Measurements VT = -215.208 +- 0.842 uVs I1Y = -43.042 +- 0.169 uTm 21:26:16 Wire Position: X0 = -1.0000 mm, Y0 = 0.0000 mm 21:26:18 Integrated Voltage Measurement... 21:26:27 Voltage Offset Correction: v_off_ini = 34.105 uV, v_off_fin = 21.148 uV 21:26:27 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -106.666 uVs 21:26:27 Integrated Voltage Measurement... 21:26:36 Voltage Offset Correction: v_off_ini = 37.389 uV, v_off_fin = 30.336 uV 21:26:36 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 92.313 uVs 21:26:36 Integrated Voltage Measurement... 21:26:45 Voltage Offset Correction: v_off_ini = 24.928 uV, v_off_fin = 10.035 uV 21:26:45 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -89.438 uVs 21:26:45 Integrated Voltage Measurement... 21:26:54 Voltage Offset Correction: v_off_ini = 22.952 uV, v_off_fin = 20.196 uV 21:26:54 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 98.697 uVs 21:26:54 Integrated Voltage Measurement... 21:27:03 Voltage Offset Correction: v_off_ini = 31.653 uV, v_off_fin = 27.498 uV 21:27:03 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -91.629 uVs 21:27:03 Integrated Voltage Measurement... 21:27:12 Voltage Offset Correction: v_off_ini = 57.447 uV, v_off_fin = 45.371 uV 21:27:12 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 83.351 uVs 21:27:15 I2X Average and RMS of Measurements VT = -93.682 +- 4.906 uVs I2X = 8.619 +- 0.451 uTm^2 21:27:20 Wire Position: X0 = -1.0000 mm, Y0 = 0.0000 mm 21:27:23 Integrated Voltage Measurement... 21:27:31 Voltage Offset Correction: v_off_ini = 26.841 uV, v_off_fin = 31.273 uV 21:27:32 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -228.861 uVs 21:27:32 Integrated Voltage Measurement... 21:27:41 Voltage Offset Correction: v_off_ini = 44.045 uV, v_off_fin = 21.706 uV 21:27:41 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 249.031 uVs 21:27:41 Integrated Voltage Measurement... 21:27:50 Voltage Offset Correction: v_off_ini = 24.934 uV, v_off_fin = 29.864 uV 21:27:50 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -235.368 uVs 21:27:50 Integrated Voltage Measurement... 21:27:59 Voltage Offset Correction: v_off_ini = 36.677 uV, v_off_fin = 36.038 uV 21:27:59 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = 247.969 uVs 21:27:59 Integrated Voltage Measurement... 21:28:08 Voltage Offset Correction: v_off_ini = 36.362 uV, v_off_fin = 33.771 uV 21:28:08 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -244.685 uVs 21:28:08 Integrated Voltage Measurement... 21:28:17 Voltage Offset Correction: v_off_ini = 29.363 uV, v_off_fin = 34.938 uV 21:28:17 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 256.127 uVs 21:28:19 I2Y Average and RMS of Measurements VT = -243.673 +- 4.889 uVs I2Y = -22.418 +- 0.450 uTm^2 21:28:24 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 21:28:28 Integrated Voltage Measurement... 21:28:37 Voltage Offset Correction: v_off_ini = 28.794 uV, v_off_fin = 25.558 uV 21:28:37 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -68.165 uVs 21:28:37 Integrated Voltage Measurement... 21:28:46 Voltage Offset Correction: v_off_ini = 42.803 uV, v_off_fin = 36.417 uV 21:28:46 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 93.276 uVs 21:28:46 Integrated Voltage Measurement... 21:28:55 Voltage Offset Correction: v_off_ini = 41.453 uV, v_off_fin = 37.415 uV 21:28:55 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -92.003 uVs 21:28:55 Integrated Voltage Measurement... 21:29:04 Voltage Offset Correction: v_off_ini = 38.203 uV, v_off_fin = 41.507 uV 21:29:04 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 101.796 uVs 21:29:04 Integrated Voltage Measurement... 21:29:13 Voltage Offset Correction: v_off_ini = 35.478 uV, v_off_fin = 33.769 uV 21:29:13 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -87.545 uVs 21:29:13 Integrated Voltage Measurement... 21:29:22 Voltage Offset Correction: v_off_ini = 33.785 uV, v_off_fin = 44.848 uV 21:29:22 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 98.379 uVs 21:29:26 I1X Average and RMS of Measurements VT = -90.194 +- 6.889 uVs I1X = 18.039 +- 1.378 uTm 21:29:29 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 21:29:33 Integrated Voltage Measurement... 21:29:42 Voltage Offset Correction: v_off_ini = 27.473 uV, v_off_fin = 31.992 uV 21:29:42 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -204.621 uVs 21:29:42 Integrated Voltage Measurement... 21:29:51 Voltage Offset Correction: v_off_ini = 35.635 uV, v_off_fin = 48.951 uV 21:29:51 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 194.795 uVs 21:29:51 Integrated Voltage Measurement... 21:30:00 Voltage Offset Correction: v_off_ini = 34.380 uV, v_off_fin = 27.721 uV 21:30:00 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -209.994 uVs 21:30:00 Integrated Voltage Measurement... 21:30:09 Voltage Offset Correction: v_off_ini = 40.695 uV, v_off_fin = 49.220 uV 21:30:09 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 194.358 uVs 21:30:09 Integrated Voltage Measurement... 21:30:18 Voltage Offset Correction: v_off_ini = 28.151 uV, v_off_fin = 23.715 uV 21:30:18 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -208.450 uVs 21:30:18 Integrated Voltage Measurement... 21:30:27 Voltage Offset Correction: v_off_ini = 31.136 uV, v_off_fin = 45.861 uV 21:30:27 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 208.061 uVs 21:30:31 I1Y Average and RMS of Measurements VT = -203.380 +- 3.592 uVs I1Y = -40.676 +- 0.718 uTm 21:30:37 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 21:30:39 Integrated Voltage Measurement... 21:30:48 Voltage Offset Correction: v_off_ini = 22.094 uV, v_off_fin = 10.481 uV 21:30:48 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -107.234 uVs 21:30:48 Integrated Voltage Measurement... 21:30:57 Voltage Offset Correction: v_off_ini = 31.553 uV, v_off_fin = 33.480 uV 21:30:57 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 113.932 uVs 21:30:57 Integrated Voltage Measurement... 21:31:06 Voltage Offset Correction: v_off_ini = 22.086 uV, v_off_fin = 21.642 uV 21:31:06 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -115.701 uVs 21:31:06 Integrated Voltage Measurement... 21:31:15 Voltage Offset Correction: v_off_ini = 16.668 uV, v_off_fin = 23.863 uV 21:31:15 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 111.413 uVs 21:31:15 Integrated Voltage Measurement... 21:31:24 Voltage Offset Correction: v_off_ini = 26.362 uV, v_off_fin = 14.372 uV 21:31:24 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -106.462 uVs 21:31:24 Integrated Voltage Measurement... 21:31:33 Voltage Offset Correction: v_off_ini = 17.929 uV, v_off_fin = 36.409 uV 21:31:33 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 92.980 uVs 21:31:36 I2X Average and RMS of Measurements VT = -107.954 +- 5.947 uVs I2X = 9.932 +- 0.547 uTm^2 21:31:41 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 21:31:43 Integrated Voltage Measurement... 21:31:52 Voltage Offset Correction: v_off_ini = 26.105 uV, v_off_fin = 24.417 uV 21:31:52 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -240.341 uVs 21:31:53 Integrated Voltage Measurement... 21:32:01 Voltage Offset Correction: v_off_ini = 18.423 uV, v_off_fin = 16.947 uV 21:32:02 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 243.085 uVs 21:32:02 Integrated Voltage Measurement... 21:32:11 Voltage Offset Correction: v_off_ini = 20.461 uV, v_off_fin = 33.851 uV 21:32:11 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -241.475 uVs 21:32:11 Integrated Voltage Measurement... 21:32:20 Voltage Offset Correction: v_off_ini = 13.667 uV, v_off_fin = 11.228 uV 21:32:20 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 246.123 uVs 21:32:20 Integrated Voltage Measurement... 21:32:29 Voltage Offset Correction: v_off_ini = 17.979 uV, v_off_fin = 18.147 uV 21:32:29 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -238.839 uVs 21:32:29 Integrated Voltage Measurement... 21:32:38 Voltage Offset Correction: v_off_ini = 18.705 uV, v_off_fin = 21.688 uV 21:32:38 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 244.198 uVs 21:32:40 I2Y Average and RMS of Measurements VT = -242.344 +- 1.032 uVs I2Y = -22.296 +- 0.095 uTm^2 21:32:45 Wire Position: X0 = 1.0000 mm, Y0 = 0.0000 mm 21:32:49 Integrated Voltage Measurement... 21:32:58 Voltage Offset Correction: v_off_ini = 33.284 uV, v_off_fin = 30.627 uV 21:32:58 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.4998 mm dXY = -0.4999 mm VT = -100.639 uVs 21:32:58 Integrated Voltage Measurement... 21:33:07 Voltage Offset Correction: v_off_ini = 27.045 uV, v_off_fin = 29.063 uV 21:33:07 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 78.537 uVs 21:33:07 Integrated Voltage Measurement... 21:33:16 Voltage Offset Correction: v_off_ini = 26.208 uV, v_off_fin = 25.712 uV 21:33:16 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -86.451 uVs 21:33:16 Integrated Voltage Measurement... 21:33:25 Voltage Offset Correction: v_off_ini = 25.556 uV, v_off_fin = 45.888 uV 21:33:25 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 83.923 uVs 21:33:25 Integrated Voltage Measurement... 21:33:34 Voltage Offset Correction: v_off_ini = 43.507 uV, v_off_fin = 30.766 uV 21:33:34 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -94.083 uVs 21:33:34 Integrated Voltage Measurement... 21:33:43 Voltage Offset Correction: v_off_ini = 36.039 uV, v_off_fin = 35.086 uV 21:33:43 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = 98.122 uVs 21:33:47 I1X Average and RMS of Measurements VT = -90.293 +- 4.484 uVs I1X = 18.060 +- 0.897 uTm 21:33:50 Wire Position: X0 = 1.0000 mm, Y0 = 0.0000 mm 21:33:54 Integrated Voltage Measurement... 21:34:03 Voltage Offset Correction: v_off_ini = 37.655 uV, v_off_fin = 37.385 uV 21:34:03 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -225.169 uVs 21:34:03 Integrated Voltage Measurement... 21:34:12 Voltage Offset Correction: v_off_ini = 32.427 uV, v_off_fin = 48.766 uV 21:34:12 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 222.180 uVs 21:34:12 Integrated Voltage Measurement... 21:34:21 Voltage Offset Correction: v_off_ini = 51.644 uV, v_off_fin = 46.658 uV 21:34:21 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -223.230 uVs 21:34:21 Integrated Voltage Measurement... 21:34:30 Voltage Offset Correction: v_off_ini = 45.353 uV, v_off_fin = 66.379 uV 21:34:30 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 205.998 uVs 21:34:30 Integrated Voltage Measurement... 21:34:39 Voltage Offset Correction: v_off_ini = 65.308 uV, v_off_fin = 52.579 uV 21:34:39 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -225.827 uVs 21:34:39 Integrated Voltage Measurement... 21:34:48 Voltage Offset Correction: v_off_ini = 59.524 uV, v_off_fin = 59.061 uV 21:34:48 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 223.629 uVs 21:34:52 I1Y Average and RMS of Measurements VT = -221.005 +- 4.540 uVs I1Y = -44.201 +- 0.908 uTm 21:34:57 Wire Position: X0 = 1.0000 mm, Y0 = 0.0000 mm 21:35:00 Integrated Voltage Measurement... 21:35:09 Voltage Offset Correction: v_off_ini = 63.090 uV, v_off_fin = 58.110 uV 21:35:09 Integrated Voltage Measurement dXYus = -1.0001 mm, dXYds = 0.0000 mm dXY = -1.0001 mm VT = -89.680 uVs 21:35:09 Integrated Voltage Measurement... 21:35:18 Voltage Offset Correction: v_off_ini = 61.047 uV, v_off_fin = 69.705 uV 21:35:18 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 98.657 uVs 21:35:18 Integrated Voltage Measurement... 21:35:27 Voltage Offset Correction: v_off_ini = 62.232 uV, v_off_fin = 52.833 uV 21:35:27 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -114.861 uVs 21:35:27 Integrated Voltage Measurement... 21:35:36 Voltage Offset Correction: v_off_ini = 55.667 uV, v_off_fin = 75.121 uV 21:35:36 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 90.358 uVs 21:35:36 Integrated Voltage Measurement... 21:35:45 Voltage Offset Correction: v_off_ini = 51.594 uV, v_off_fin = 50.933 uV 21:35:45 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -90.070 uVs 21:35:45 Integrated Voltage Measurement... 21:35:54 Voltage Offset Correction: v_off_ini = 56.957 uV, v_off_fin = 73.518 uV 21:35:54 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 91.232 uVs 21:35:57 I2X Average and RMS of Measurements VT = -95.810 +- 5.018 uVs I2X = 8.815 +- 0.462 uTm^2 21:36:02 Wire Position: X0 = 1.0000 mm, Y0 = 0.0000 mm 21:36:04 Integrated Voltage Measurement... 21:36:13 Voltage Offset Correction: v_off_ini = 74.841 uV, v_off_fin = 65.822 uV 21:36:13 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -247.347 uVs 21:36:13 Integrated Voltage Measurement... 21:36:22 Voltage Offset Correction: v_off_ini = 66.881 uV, v_off_fin = 60.380 uV 21:36:22 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 245.313 uVs 21:36:22 Integrated Voltage Measurement... 21:36:31 Voltage Offset Correction: v_off_ini = 63.533 uV, v_off_fin = 72.166 uV 21:36:31 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -250.305 uVs 21:36:31 Integrated Voltage Measurement... 21:36:40 Voltage Offset Correction: v_off_ini = 72.014 uV, v_off_fin = 73.225 uV 21:36:41 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 229.357 uVs 21:36:41 Integrated Voltage Measurement... 21:36:50 Voltage Offset Correction: v_off_ini = 85.024 uV, v_off_fin = 64.863 uV 21:36:50 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = -260.471 uVs 21:36:50 Integrated Voltage Measurement... 21:36:59 Voltage Offset Correction: v_off_ini = 71.081 uV, v_off_fin = 68.241 uV 21:36:59 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 251.829 uVs 21:37:01 I2Y Average and RMS of Measurements VT = -247.437 +- 6.708 uVs I2Y = -22.765 +- 0.618 uTm^2 21:37:06 Wire Position: X0 = 2.0000 mm, Y0 = 0.0000 mm 21:37:10 Integrated Voltage Measurement... 21:37:19 Voltage Offset Correction: v_off_ini = 74.081 uV, v_off_fin = 71.679 uV 21:37:19 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -88.583 uVs 21:37:19 Integrated Voltage Measurement... 21:37:28 Voltage Offset Correction: v_off_ini = 60.352 uV, v_off_fin = 62.694 uV 21:37:28 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 88.478 uVs 21:37:28 Integrated Voltage Measurement... 21:37:37 Voltage Offset Correction: v_off_ini = 59.193 uV, v_off_fin = 59.222 uV 21:37:37 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -84.597 uVs 21:37:37 Integrated Voltage Measurement... 21:37:46 Voltage Offset Correction: v_off_ini = 60.430 uV, v_off_fin = 63.197 uV 21:37:46 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 103.399 uVs 21:37:46 Integrated Voltage Measurement... 21:37:55 Voltage Offset Correction: v_off_ini = 60.043 uV, v_off_fin = 64.333 uV 21:37:55 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -100.980 uVs 21:37:55 Integrated Voltage Measurement... 21:38:04 Voltage Offset Correction: v_off_ini = 58.019 uV, v_off_fin = 64.861 uV 21:38:04 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 109.034 uVs 21:38:08 I1X Average and RMS of Measurements VT = -95.845 +- 6.852 uVs I1X = 19.169 +- 1.370 uTm 21:38:11 Wire Position: X0 = 2.0000 mm, Y0 = 0.0000 mm 21:38:15 Integrated Voltage Measurement... 21:38:24 Voltage Offset Correction: v_off_ini = 50.031 uV, v_off_fin = 40.972 uV 21:38:24 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.4999 mm dXY = 0.4999 mm VT = -208.236 uVs 21:38:24 Integrated Voltage Measurement... 21:38:33 Voltage Offset Correction: v_off_ini = 59.934 uV, v_off_fin = 70.731 uV 21:38:33 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 216.745 uVs 21:38:33 Integrated Voltage Measurement... 21:38:42 Voltage Offset Correction: v_off_ini = 65.493 uV, v_off_fin = 59.105 uV 21:38:42 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -212.183 uVs 21:38:42 Integrated Voltage Measurement... 21:38:51 Voltage Offset Correction: v_off_ini = 59.655 uV, v_off_fin = 79.442 uV 21:38:51 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 203.155 uVs 21:38:51 Integrated Voltage Measurement... 21:39:00 Voltage Offset Correction: v_off_ini = 57.695 uV, v_off_fin = 48.229 uV 21:39:00 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -201.016 uVs 21:39:00 Integrated Voltage Measurement... 21:39:09 Voltage Offset Correction: v_off_ini = 73.399 uV, v_off_fin = 88.317 uV 21:39:09 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 196.899 uVs 21:39:13 I1Y Average and RMS of Measurements VT = -206.372 +- 5.600 uVs I1Y = -41.276 +- 1.122 uTm 21:39:18 Wire Position: X0 = 2.0000 mm, Y0 = 0.0000 mm 21:39:21 Integrated Voltage Measurement... 21:39:30 Voltage Offset Correction: v_off_ini = 68.107 uV, v_off_fin = 49.689 uV 21:39:30 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -93.006 uVs 21:39:30 Integrated Voltage Measurement... 21:39:39 Voltage Offset Correction: v_off_ini = 54.878 uV, v_off_fin = 55.398 uV 21:39:39 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 117.595 uVs 21:39:39 Integrated Voltage Measurement... 21:39:48 Voltage Offset Correction: v_off_ini = 56.551 uV, v_off_fin = 38.450 uV 21:39:48 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -101.154 uVs 21:39:48 Integrated Voltage Measurement... 21:39:57 Voltage Offset Correction: v_off_ini = 54.152 uV, v_off_fin = 61.886 uV 21:39:57 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 99.257 uVs 21:39:57 Integrated Voltage Measurement... 21:40:06 Voltage Offset Correction: v_off_ini = 42.154 uV, v_off_fin = 19.638 uV 21:40:06 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -93.180 uVs 21:40:06 Integrated Voltage Measurement... 21:40:15 Voltage Offset Correction: v_off_ini = 30.519 uV, v_off_fin = 45.452 uV 21:40:15 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 112.555 uVs 21:40:17 I2X Average and RMS of Measurements VT = -102.791 +- 2.081 uVs I2X = 9.457 +- 0.191 uTm^2 21:40:22 Wire Position: X0 = 2.0000 mm, Y0 = 0.0000 mm 21:40:25 Integrated Voltage Measurement... 21:40:34 Voltage Offset Correction: v_off_ini = 38.265 uV, v_off_fin = 44.300 uV 21:40:34 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -256.856 uVs 21:40:34 Integrated Voltage Measurement... 21:40:43 Voltage Offset Correction: v_off_ini = 32.745 uV, v_off_fin = 30.378 uV 21:40:43 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 255.696 uVs 21:40:43 Integrated Voltage Measurement... 21:40:52 Voltage Offset Correction: v_off_ini = 30.819 uV, v_off_fin = 31.172 uV 21:40:52 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -249.618 uVs 21:40:52 Integrated Voltage Measurement... 21:41:01 Voltage Offset Correction: v_off_ini = 43.238 uV, v_off_fin = 49.386 uV 21:41:01 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 249.571 uVs 21:41:01 Integrated Voltage Measurement... 21:41:10 Voltage Offset Correction: v_off_ini = 24.757 uV, v_off_fin = 39.555 uV 21:41:10 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -242.073 uVs 21:41:10 Integrated Voltage Measurement... 21:41:19 Voltage Offset Correction: v_off_ini = 34.274 uV, v_off_fin = 35.502 uV 21:41:19 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 237.387 uVs 21:41:21 I2Y Average and RMS of Measurements VT = -248.534 +- 6.796 uVs I2Y = -22.865 +- 0.625 uTm^2 21:41:29 Wire Position: X0 = 0.0000 mm, Y0 = -1.0000 mm 21:41:33 Integrated Voltage Measurement... 21:41:42 Voltage Offset Correction: v_off_ini = 56.346 uV, v_off_fin = 40.017 uV 21:41:42 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -95.909 uVs 21:41:42 Integrated Voltage Measurement... 21:41:51 Voltage Offset Correction: v_off_ini = 37.481 uV, v_off_fin = 35.291 uV 21:41:51 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 127.236 uVs 21:41:51 Integrated Voltage Measurement... 21:42:00 Voltage Offset Correction: v_off_ini = 29.813 uV, v_off_fin = 24.695 uV 21:42:00 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -92.650 uVs 21:42:00 Integrated Voltage Measurement... 21:42:09 Voltage Offset Correction: v_off_ini = 18.821 uV, v_off_fin = 19.930 uV 21:42:09 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 78.765 uVs 21:42:09 Integrated Voltage Measurement... 21:42:18 Voltage Offset Correction: v_off_ini = 13.876 uV, v_off_fin = 10.152 uV 21:42:18 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -97.381 uVs 21:42:18 Integrated Voltage Measurement... 21:42:27 Voltage Offset Correction: v_off_ini = 20.578 uV, v_off_fin = 31.102 uV 21:42:27 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 78.926 uVs 21:42:32 I1X Average and RMS of Measurements VT = -95.144 +- 11.659 uVs I1X = 19.029 +- 2.332 uTm 21:42:34 Wire Position: X0 = 0.0000 mm, Y0 = -1.0000 mm 21:42:38 Integrated Voltage Measurement... 21:42:47 Voltage Offset Correction: v_off_ini = 37.337 uV, v_off_fin = 20.852 uV 21:42:47 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.4999 mm dXY = 0.4999 mm VT = -195.197 uVs 21:42:47 Integrated Voltage Measurement... 21:42:56 Voltage Offset Correction: v_off_ini = 44.326 uV, v_off_fin = 39.728 uV 21:42:56 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 193.217 uVs 21:42:56 Integrated Voltage Measurement... 21:43:05 Voltage Offset Correction: v_off_ini = 22.123 uV, v_off_fin = 23.614 uV 21:43:05 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -211.783 uVs 21:43:05 Integrated Voltage Measurement... 21:43:14 Voltage Offset Correction: v_off_ini = 33.175 uV, v_off_fin = 48.346 uV 21:43:14 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 193.918 uVs 21:43:14 Integrated Voltage Measurement... 21:43:23 Voltage Offset Correction: v_off_ini = 31.426 uV, v_off_fin = 23.498 uV 21:43:23 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -205.509 uVs 21:43:23 Integrated Voltage Measurement... 21:43:32 Voltage Offset Correction: v_off_ini = 34.775 uV, v_off_fin = 45.187 uV 21:43:32 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 210.818 uVs 21:43:36 I1Y Average and RMS of Measurements VT = -201.740 +- 5.752 uVs I1Y = -40.349 +- 1.149 uTm 21:43:39 Wire Position: X0 = 0.0000 mm, Y0 = -1.0000 mm 21:43:42 Integrated Voltage Measurement... 21:43:51 Voltage Offset Correction: v_off_ini = 20.520 uV, v_off_fin = 24.671 uV 21:43:51 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -81.434 uVs 21:43:51 Integrated Voltage Measurement... 21:44:00 Voltage Offset Correction: v_off_ini = 40.301 uV, v_off_fin = 43.754 uV 21:44:00 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 95.556 uVs 21:44:00 Integrated Voltage Measurement... 21:44:09 Voltage Offset Correction: v_off_ini = 41.668 uV, v_off_fin = 27.977 uV 21:44:09 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -89.755 uVs 21:44:09 Integrated Voltage Measurement... 21:44:18 Voltage Offset Correction: v_off_ini = 34.730 uV, v_off_fin = 51.383 uV 21:44:18 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 84.394 uVs 21:44:18 Integrated Voltage Measurement... 21:44:27 Voltage Offset Correction: v_off_ini = 34.627 uV, v_off_fin = 33.950 uV 21:44:27 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -86.330 uVs 21:44:27 Integrated Voltage Measurement... 21:44:36 Voltage Offset Correction: v_off_ini = 41.331 uV, v_off_fin = 45.067 uV 21:44:36 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 72.788 uVs 21:44:38 I2X Average and RMS of Measurements VT = -85.043 +- 3.921 uVs I2X = 7.824 +- 0.361 uTm^2 21:44:41 Wire Position: X0 = 0.0000 mm, Y0 = -1.0000 mm 21:44:43 Integrated Voltage Measurement... 21:44:52 Voltage Offset Correction: v_off_ini = 39.984 uV, v_off_fin = 34.726 uV 21:44:52 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -254.542 uVs 21:44:52 Integrated Voltage Measurement... 21:45:01 Voltage Offset Correction: v_off_ini = 45.926 uV, v_off_fin = 38.210 uV 21:45:01 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 234.519 uVs 21:45:01 Integrated Voltage Measurement... 21:45:10 Voltage Offset Correction: v_off_ini = 38.341 uV, v_off_fin = 37.326 uV 21:45:10 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -249.498 uVs 21:45:10 Integrated Voltage Measurement... 21:45:19 Voltage Offset Correction: v_off_ini = 32.974 uV, v_off_fin = 30.733 uV 21:45:19 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 248.132 uVs 21:45:19 Integrated Voltage Measurement... 21:45:28 Voltage Offset Correction: v_off_ini = 34.858 uV, v_off_fin = 40.145 uV 21:45:28 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -237.911 uVs 21:45:28 Integrated Voltage Measurement... 21:45:37 Voltage Offset Correction: v_off_ini = 37.848 uV, v_off_fin = 28.625 uV 21:45:37 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 245.073 uVs 21:45:40 I2Y Average and RMS of Measurements VT = -244.946 +- 3.004 uVs I2Y = -22.535 +- 0.276 uTm^2 21:45:44 Wire Position: X0 = 0.0000 mm, Y0 = -0.5000 mm 21:45:48 Integrated Voltage Measurement... 21:45:57 Voltage Offset Correction: v_off_ini = 37.420 uV, v_off_fin = 40.829 uV 21:45:57 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -82.975 uVs 21:45:57 Integrated Voltage Measurement... 21:46:06 Voltage Offset Correction: v_off_ini = 41.068 uV, v_off_fin = 40.291 uV 21:46:06 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 84.089 uVs 21:46:06 Integrated Voltage Measurement... 21:46:15 Voltage Offset Correction: v_off_ini = 35.850 uV, v_off_fin = 36.590 uV 21:46:15 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -95.300 uVs 21:46:15 Integrated Voltage Measurement... 21:46:24 Voltage Offset Correction: v_off_ini = 31.612 uV, v_off_fin = 35.257 uV 21:46:24 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 80.717 uVs 21:46:24 Integrated Voltage Measurement... 21:46:33 Voltage Offset Correction: v_off_ini = 41.741 uV, v_off_fin = 40.905 uV 21:46:33 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -85.063 uVs 21:46:33 Integrated Voltage Measurement... 21:46:42 Voltage Offset Correction: v_off_ini = 41.031 uV, v_off_fin = 48.695 uV 21:46:42 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 84.712 uVs 21:46:46 I1X Average and RMS of Measurements VT = -85.476 +- 1.874 uVs I1X = 17.095 +- 0.375 uTm 21:46:49 Wire Position: X0 = 0.0000 mm, Y0 = -0.5000 mm 21:46:53 Integrated Voltage Measurement... 21:47:02 Voltage Offset Correction: v_off_ini = 61.719 uV, v_off_fin = 52.923 uV 21:47:02 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -206.503 uVs 21:47:02 Integrated Voltage Measurement... 21:47:11 Voltage Offset Correction: v_off_ini = 54.222 uV, v_off_fin = 67.173 uV 21:47:11 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = 218.302 uVs 21:47:11 Integrated Voltage Measurement... 21:47:20 Voltage Offset Correction: v_off_ini = 57.629 uV, v_off_fin = 49.693 uV 21:47:20 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -196.059 uVs 21:47:20 Integrated Voltage Measurement... 21:47:29 Voltage Offset Correction: v_off_ini = 56.401 uV, v_off_fin = 68.924 uV 21:47:29 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 190.793 uVs 21:47:29 Integrated Voltage Measurement... 21:47:38 Voltage Offset Correction: v_off_ini = 61.918 uV, v_off_fin = 53.424 uV 21:47:38 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -215.309 uVs 21:47:38 Integrated Voltage Measurement... 21:47:47 Voltage Offset Correction: v_off_ini = 55.963 uV, v_off_fin = 71.840 uV 21:47:47 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 196.497 uVs 21:47:51 I1Y Average and RMS of Measurements VT = -203.910 +- 7.874 uVs I1Y = -40.783 +- 1.576 uTm 21:47:54 Wire Position: X0 = 0.0000 mm, Y0 = -0.5000 mm 21:47:56 Integrated Voltage Measurement... 21:48:05 Voltage Offset Correction: v_off_ini = 69.562 uV, v_off_fin = 51.844 uV 21:48:05 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -108.120 uVs 21:48:05 Integrated Voltage Measurement... 21:48:14 Voltage Offset Correction: v_off_ini = 66.403 uV, v_off_fin = 78.794 uV 21:48:14 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 85.937 uVs 21:48:14 Integrated Voltage Measurement... 21:48:23 Voltage Offset Correction: v_off_ini = 56.064 uV, v_off_fin = 54.223 uV 21:48:23 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -88.346 uVs 21:48:23 Integrated Voltage Measurement... 21:48:32 Voltage Offset Correction: v_off_ini = 58.052 uV, v_off_fin = 62.960 uV 21:48:32 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 75.677 uVs 21:48:32 Integrated Voltage Measurement... 21:48:41 Voltage Offset Correction: v_off_ini = 46.285 uV, v_off_fin = 47.106 uV 21:48:41 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -93.749 uVs 21:48:41 Integrated Voltage Measurement... 21:48:50 Voltage Offset Correction: v_off_ini = 51.278 uV, v_off_fin = 64.028 uV 21:48:50 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 90.300 uVs 21:48:53 I2X Average and RMS of Measurements VT = -90.355 +- 6.243 uVs I2X = 8.313 +- 0.574 uTm^2 21:48:55 Wire Position: X0 = 0.0000 mm, Y0 = -0.5000 mm 21:48:58 Integrated Voltage Measurement... 21:49:07 Voltage Offset Correction: v_off_ini = 67.939 uV, v_off_fin = 64.789 uV 21:49:07 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -245.291 uVs 21:49:07 Integrated Voltage Measurement... 21:49:16 Voltage Offset Correction: v_off_ini = 69.994 uV, v_off_fin = 74.453 uV 21:49:16 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 252.584 uVs 21:49:16 Integrated Voltage Measurement... 21:49:25 Voltage Offset Correction: v_off_ini = 71.641 uV, v_off_fin = 83.182 uV 21:49:25 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = -239.076 uVs 21:49:25 Integrated Voltage Measurement... 21:49:34 Voltage Offset Correction: v_off_ini = 80.954 uV, v_off_fin = 61.795 uV 21:49:34 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 244.332 uVs 21:49:34 Integrated Voltage Measurement... 21:49:43 Voltage Offset Correction: v_off_ini = 62.056 uV, v_off_fin = 66.276 uV 21:49:43 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -240.904 uVs 21:49:43 Integrated Voltage Measurement... 21:49:52 Voltage Offset Correction: v_off_ini = 66.571 uV, v_off_fin = 70.058 uV 21:49:52 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 238.338 uVs 21:49:54 I2Y Average and RMS of Measurements VT = -243.421 +- 3.993 uVs I2Y = -22.395 +- 0.367 uTm^2 21:49:58 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 21:50:03 Integrated Voltage Measurement... 21:50:12 Voltage Offset Correction: v_off_ini = 71.168 uV, v_off_fin = 59.405 uV 21:50:12 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.4998 mm dXY = -0.4999 mm VT = -78.344 uVs 21:50:12 Integrated Voltage Measurement... 21:50:21 Voltage Offset Correction: v_off_ini = 56.815 uV, v_off_fin = 70.209 uV 21:50:21 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 118.282 uVs 21:50:21 Integrated Voltage Measurement... 21:50:30 Voltage Offset Correction: v_off_ini = 74.596 uV, v_off_fin = 57.474 uV 21:50:30 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -97.647 uVs 21:50:30 Integrated Voltage Measurement... 21:50:39 Voltage Offset Correction: v_off_ini = 72.274 uV, v_off_fin = 65.547 uV 21:50:39 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 90.726 uVs 21:50:39 Integrated Voltage Measurement... 21:50:48 Voltage Offset Correction: v_off_ini = 54.636 uV, v_off_fin = 65.345 uV 21:50:48 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -81.847 uVs 21:50:48 Integrated Voltage Measurement... 21:50:57 Voltage Offset Correction: v_off_ini = 58.537 uV, v_off_fin = 66.578 uV 21:50:57 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 118.184 uVs 21:51:01 I1X Average and RMS of Measurements VT = -97.505 +- 2.447 uVs I1X = 19.502 +- 0.490 uTm 21:51:03 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 21:51:08 Integrated Voltage Measurement... 21:51:17 Voltage Offset Correction: v_off_ini = 80.015 uV, v_off_fin = 65.813 uV 21:51:17 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -209.081 uVs 21:51:17 Integrated Voltage Measurement... 21:51:26 Voltage Offset Correction: v_off_ini = 74.791 uV, v_off_fin = 95.569 uV 21:51:26 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 204.358 uVs 21:51:26 Integrated Voltage Measurement... 21:51:35 Voltage Offset Correction: v_off_ini = 95.769 uV, v_off_fin = 75.899 uV 21:51:35 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -220.802 uVs 21:51:35 Integrated Voltage Measurement... 21:51:44 Voltage Offset Correction: v_off_ini = 75.602 uV, v_off_fin = 91.333 uV 21:51:44 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 209.678 uVs 21:51:44 Integrated Voltage Measurement... 21:51:53 Voltage Offset Correction: v_off_ini = 83.799 uV, v_off_fin = 84.918 uV 21:51:53 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -206.896 uVs 21:51:53 Integrated Voltage Measurement... 21:52:02 Voltage Offset Correction: v_off_ini = 75.658 uV, v_off_fin = 90.407 uV 21:52:02 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 217.899 uVs 21:52:06 I1Y Average and RMS of Measurements VT = -211.452 +- 3.542 uVs I1Y = -42.291 +- 0.709 uTm 21:52:08 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 21:52:11 Integrated Voltage Measurement... 21:52:20 Voltage Offset Correction: v_off_ini = 83.385 uV, v_off_fin = 75.405 uV 21:52:20 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -106.692 uVs 21:52:20 Integrated Voltage Measurement... 21:52:29 Voltage Offset Correction: v_off_ini = 86.150 uV, v_off_fin = 101.004 uV 21:52:29 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 108.623 uVs 21:52:29 Integrated Voltage Measurement... 21:52:38 Voltage Offset Correction: v_off_ini = 85.209 uV, v_off_fin = 83.133 uV 21:52:38 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -99.904 uVs 21:52:38 Integrated Voltage Measurement... 21:52:47 Voltage Offset Correction: v_off_ini = 89.551 uV, v_off_fin = 92.338 uV 21:52:47 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 107.101 uVs 21:52:47 Integrated Voltage Measurement... 21:52:56 Voltage Offset Correction: v_off_ini = 79.902 uV, v_off_fin = 77.571 uV 21:52:56 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -84.789 uVs 21:52:56 Integrated Voltage Measurement... 21:53:05 Voltage Offset Correction: v_off_ini = 83.679 uV, v_off_fin = 95.105 uV 21:53:05 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 79.125 uVs 21:53:07 I2X Average and RMS of Measurements VT = -97.706 +- 11.264 uVs I2X = 8.989 +- 1.036 uTm^2 21:53:10 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 21:53:12 Integrated Voltage Measurement... 21:53:21 Voltage Offset Correction: v_off_ini = 89.743 uV, v_off_fin = 82.730 uV 21:53:21 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -245.866 uVs 21:53:22 Integrated Voltage Measurement... 21:53:30 Voltage Offset Correction: v_off_ini = 94.567 uV, v_off_fin = 89.245 uV 21:53:31 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 255.581 uVs 21:53:31 Integrated Voltage Measurement... 21:53:40 Voltage Offset Correction: v_off_ini = 81.784 uV, v_off_fin = 78.182 uV 21:53:40 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -260.613 uVs 21:53:40 Integrated Voltage Measurement... 21:53:49 Voltage Offset Correction: v_off_ini = 77.993 uV, v_off_fin = 74.536 uV 21:53:49 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 252.662 uVs 21:53:49 Integrated Voltage Measurement... 21:53:58 Voltage Offset Correction: v_off_ini = 74.785 uV, v_off_fin = 72.450 uV 21:53:58 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -241.262 uVs 21:53:58 Integrated Voltage Measurement... 21:54:07 Voltage Offset Correction: v_off_ini = 65.980 uV, v_off_fin = 74.017 uV 21:54:07 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 246.687 uVs 21:54:09 I2Y Average and RMS of Measurements VT = -250.445 +- 5.173 uVs I2Y = -23.041 +- 0.476 uTm^2 21:54:13 Wire Position: X0 = 0.0000 mm, Y0 = 0.5000 mm 21:54:17 Integrated Voltage Measurement... 21:54:26 Voltage Offset Correction: v_off_ini = 80.716 uV, v_off_fin = 69.246 uV 21:54:26 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -108.551 uVs 21:54:26 Integrated Voltage Measurement... 21:54:35 Voltage Offset Correction: v_off_ini = 91.053 uV, v_off_fin = 76.203 uV 21:54:35 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 71.445 uVs 21:54:35 Integrated Voltage Measurement... 21:54:44 Voltage Offset Correction: v_off_ini = 79.653 uV, v_off_fin = 63.020 uV 21:54:44 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -100.768 uVs 21:54:44 Integrated Voltage Measurement... 21:54:53 Voltage Offset Correction: v_off_ini = 80.445 uV, v_off_fin = 85.703 uV 21:54:53 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 81.215 uVs 21:54:53 Integrated Voltage Measurement... 21:55:02 Voltage Offset Correction: v_off_ini = 80.621 uV, v_off_fin = 70.549 uV 21:55:02 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -88.892 uVs 21:55:02 Integrated Voltage Measurement... 21:55:11 Voltage Offset Correction: v_off_ini = 84.971 uV, v_off_fin = 87.030 uV 21:55:11 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 100.927 uVs 21:55:16 I1X Average and RMS of Measurements VT = -91.966 +- 2.120 uVs I1X = 18.393 +- 0.424 uTm 21:55:18 Wire Position: X0 = 0.0000 mm, Y0 = 0.5000 mm 21:55:22 Integrated Voltage Measurement... 21:55:31 Voltage Offset Correction: v_off_ini = 86.907 uV, v_off_fin = 81.400 uV 21:55:31 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -216.676 uVs 21:55:31 Integrated Voltage Measurement... 21:55:40 Voltage Offset Correction: v_off_ini = 93.836 uV, v_off_fin = 112.465 uV 21:55:40 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 201.174 uVs 21:55:40 Integrated Voltage Measurement... 21:55:49 Voltage Offset Correction: v_off_ini = 96.354 uV, v_off_fin = 82.453 uV 21:55:49 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -209.852 uVs 21:55:49 Integrated Voltage Measurement... 21:55:58 Voltage Offset Correction: v_off_ini = 85.873 uV, v_off_fin = 97.482 uV 21:55:58 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 210.641 uVs 21:55:58 Integrated Voltage Measurement... 21:56:07 Voltage Offset Correction: v_off_ini = 85.774 uV, v_off_fin = 76.576 uV 21:56:07 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -187.440 uVs 21:56:07 Integrated Voltage Measurement... 21:56:16 Voltage Offset Correction: v_off_ini = 90.115 uV, v_off_fin = 88.448 uV 21:56:16 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 215.719 uVs 21:56:21 I1Y Average and RMS of Measurements VT = -206.917 +- 3.813 uVs I1Y = -41.382 +- 0.763 uTm 21:56:23 Wire Position: X0 = 0.0000 mm, Y0 = 0.5000 mm 21:56:26 Integrated Voltage Measurement... 21:56:35 Voltage Offset Correction: v_off_ini = 70.557 uV, v_off_fin = 81.386 uV 21:56:35 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -100.030 uVs 21:56:35 Integrated Voltage Measurement... 21:56:44 Voltage Offset Correction: v_off_ini = 79.616 uV, v_off_fin = 83.479 uV 21:56:44 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 75.051 uVs 21:56:44 Integrated Voltage Measurement... 21:56:53 Voltage Offset Correction: v_off_ini = 82.798 uV, v_off_fin = 74.227 uV 21:56:53 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -106.372 uVs 21:56:53 Integrated Voltage Measurement... 21:57:02 Voltage Offset Correction: v_off_ini = 81.553 uV, v_off_fin = 77.303 uV 21:57:02 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 90.356 uVs 21:57:02 Integrated Voltage Measurement... 21:57:11 Voltage Offset Correction: v_off_ini = 74.284 uV, v_off_fin = 66.034 uV 21:57:11 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -96.394 uVs 21:57:11 Integrated Voltage Measurement... 21:57:20 Voltage Offset Correction: v_off_ini = 76.441 uV, v_off_fin = 86.231 uV 21:57:20 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 92.335 uVs 21:57:22 I2X Average and RMS of Measurements VT = -93.423 +- 4.469 uVs I2X = 8.595 +- 0.411 uTm^2 21:57:25 Wire Position: X0 = 0.0000 mm, Y0 = 0.5000 mm 21:57:27 Integrated Voltage Measurement... 21:57:36 Voltage Offset Correction: v_off_ini = 79.441 uV, v_off_fin = 90.381 uV 21:57:36 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -252.199 uVs 21:57:37 Integrated Voltage Measurement... 21:57:46 Voltage Offset Correction: v_off_ini = 82.331 uV, v_off_fin = 83.759 uV 21:57:46 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 237.598 uVs 21:57:46 Integrated Voltage Measurement... 21:57:55 Voltage Offset Correction: v_off_ini = 83.054 uV, v_off_fin = 77.391 uV 21:57:55 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -258.931 uVs 21:57:55 Integrated Voltage Measurement... 21:58:04 Voltage Offset Correction: v_off_ini = 91.671 uV, v_off_fin = 92.747 uV 21:58:04 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 248.440 uVs 21:58:04 Integrated Voltage Measurement... 21:58:13 Voltage Offset Correction: v_off_ini = 104.224 uV, v_off_fin = 95.169 uV 21:58:13 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -264.621 uVs 21:58:13 Integrated Voltage Measurement... 21:58:22 Voltage Offset Correction: v_off_ini = 94.133 uV, v_off_fin = 95.145 uV 21:58:22 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = 242.129 uVs 21:58:24 I2Y Average and RMS of Measurements VT = -250.653 +- 4.071 uVs I2Y = -23.060 +- 0.375 uTm^2 21:58:28 Wire Position: X0 = 0.0000 mm, Y0 = 1.0000 mm 21:58:32 Integrated Voltage Measurement... 21:58:41 Voltage Offset Correction: v_off_ini = 102.468 uV, v_off_fin = 92.778 uV 21:58:41 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -84.548 uVs 21:58:41 Integrated Voltage Measurement... 21:58:50 Voltage Offset Correction: v_off_ini = 92.304 uV, v_off_fin = 95.746 uV 21:58:50 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 84.076 uVs 21:58:50 Integrated Voltage Measurement... 21:58:59 Voltage Offset Correction: v_off_ini = 110.451 uV, v_off_fin = 90.629 uV 21:58:59 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -96.946 uVs 21:59:00 Integrated Voltage Measurement... 21:59:08 Voltage Offset Correction: v_off_ini = 96.943 uV, v_off_fin = 98.438 uV 21:59:08 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 89.644 uVs 21:59:09 Integrated Voltage Measurement... 21:59:17 Voltage Offset Correction: v_off_ini = 107.066 uV, v_off_fin = 100.382 uV 21:59:18 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -86.403 uVs 21:59:18 Integrated Voltage Measurement... 21:59:27 Voltage Offset Correction: v_off_ini = 95.897 uV, v_off_fin = 107.558 uV 21:59:27 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 87.155 uVs 21:59:31 I1X Average and RMS of Measurements VT = -88.129 +- 3.789 uVs I1X = 17.626 +- 0.758 uTm 21:59:33 Wire Position: X0 = 0.0000 mm, Y0 = 1.0000 mm 21:59:38 Integrated Voltage Measurement... 21:59:46 Voltage Offset Correction: v_off_ini = 109.379 uV, v_off_fin = 87.705 uV 21:59:47 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -220.902 uVs 21:59:47 Integrated Voltage Measurement... 21:59:55 Voltage Offset Correction: v_off_ini = 113.121 uV, v_off_fin = 116.002 uV 21:59:55 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 220.642 uVs 21:59:56 Integrated Voltage Measurement... 22:00:04 Voltage Offset Correction: v_off_ini = 101.412 uV, v_off_fin = 77.814 uV 22:00:05 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -202.487 uVs 22:00:05 Integrated Voltage Measurement... 22:00:13 Voltage Offset Correction: v_off_ini = 88.346 uV, v_off_fin = 103.696 uV 22:00:13 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 216.081 uVs 22:00:14 Integrated Voltage Measurement... 22:00:22 Voltage Offset Correction: v_off_ini = 94.152 uV, v_off_fin = 87.676 uV 22:00:23 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -201.722 uVs 22:00:23 Integrated Voltage Measurement... 22:00:31 Voltage Offset Correction: v_off_ini = 110.142 uV, v_off_fin = 94.721 uV 22:00:32 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 217.632 uVs 22:00:36 I1Y Average and RMS of Measurements VT = -213.244 +- 5.326 uVs I1Y = -42.646 +- 1.067 uTm 22:00:38 Wire Position: X0 = 0.0000 mm, Y0 = 1.0000 mm 22:00:41 Integrated Voltage Measurement... 22:00:50 Voltage Offset Correction: v_off_ini = 95.802 uV, v_off_fin = 96.428 uV 22:00:50 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -77.153 uVs 22:00:50 Integrated Voltage Measurement... 22:00:59 Voltage Offset Correction: v_off_ini = 97.190 uV, v_off_fin = 111.990 uV 22:00:59 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 102.246 uVs 22:00:59 Integrated Voltage Measurement... 22:01:08 Voltage Offset Correction: v_off_ini = 109.348 uV, v_off_fin = 99.342 uV 22:01:08 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -98.892 uVs 22:01:08 Integrated Voltage Measurement... 22:01:17 Voltage Offset Correction: v_off_ini = 97.982 uV, v_off_fin = 109.393 uV 22:01:17 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 85.394 uVs 22:01:17 Integrated Voltage Measurement... 22:01:26 Voltage Offset Correction: v_off_ini = 89.876 uV, v_off_fin = 88.024 uV 22:01:26 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -71.761 uVs 22:01:26 Integrated Voltage Measurement... 22:01:35 Voltage Offset Correction: v_off_ini = 95.680 uV, v_off_fin = 102.312 uV 22:01:35 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 100.859 uVs 22:01:37 I2X Average and RMS of Measurements VT = -89.384 +- 2.392 uVs I2X = 8.223 +- 0.220 uTm^2 22:01:40 Wire Position: X0 = 0.0000 mm, Y0 = 1.0000 mm 22:01:43 Integrated Voltage Measurement... 22:01:52 Voltage Offset Correction: v_off_ini = 96.323 uV, v_off_fin = 98.588 uV 22:01:52 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -246.149 uVs 22:01:52 Integrated Voltage Measurement... 22:02:01 Voltage Offset Correction: v_off_ini = 88.559 uV, v_off_fin = 85.336 uV 22:02:01 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 266.461 uVs 22:02:01 Integrated Voltage Measurement... 22:02:10 Voltage Offset Correction: v_off_ini = 89.629 uV, v_off_fin = 92.795 uV 22:02:10 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -249.371 uVs 22:02:10 Integrated Voltage Measurement... 22:02:19 Voltage Offset Correction: v_off_ini = 77.408 uV, v_off_fin = 89.755 uV 22:02:19 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 240.243 uVs 22:02:19 Integrated Voltage Measurement... 22:02:28 Voltage Offset Correction: v_off_ini = 79.878 uV, v_off_fin = 81.317 uV 22:02:28 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -238.700 uVs 22:02:28 Integrated Voltage Measurement... 22:02:37 Voltage Offset Correction: v_off_ini = 71.782 uV, v_off_fin = 74.121 uV 22:02:37 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 241.586 uVs 22:02:39 I2Y Average and RMS of Measurements VT = -247.085 +- 6.792 uVs I2Y = -22.732 +- 0.625 uTm^2 22:02:44 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:05:07 Wire Position: X0 = 50.0000 mm, Y0 = 0.0000 mm