SLAC Magnetic Measurements Date: 04-23-2025 Time: 08:36:15 Project: LCLS-II-HE Magnet Type: Phase Shifter Magnet Name: HE-SXPS_16348 Dataset Number: 1 Run Purpose: Final Measurements Run Type: Stretched Wire Run Number: 1 Measurement Device: Good Field Region Operator: YL Comment: 08:36:23 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 08:36:25 Motion Parameters, X Controller, Axis 1 acc = 0.001000 m/s/s dcl = 0.001000 m/s/s vel = 0.001000 m/s pos = 0.000000 m kp = 188.600000 ki = 820.000000 kd = 491.400000 db = 0.000000 08:36:27 Motion Parameters, X Controller, Axis 2 acc = 0.001000 m/s/s dcl = 0.001000 m/s/s vel = 0.001000 m/s pos = 0.000000 m kp = 188.600000 ki = 820.000000 kd = 491.400000 db = 0.000000 08:36:28 Motion Parameters, Y Controller, Axis 1 acc = 0.001000 m/s/s dcl = 0.001000 m/s/s vel = 0.001000 m/s pos = 0.000000 m kp = 188.600000 ki = 820.000000 kd = 491.400000 db = 0.000000 08:36:30 Motion Parameters, Y Controller, Axis 2 acc = 0.001000 m/s/s dcl = 0.001000 m/s/s vel = 0.001000 m/s pos = 0.000000 m kp = 188.600000 ki = 820.000000 kd = 491.400000 db = 0.000000 08:37:07 Wire Position: X0 = -2.0000 mm, Y0 = 0.0000 mm 08:37:11 Integrated Voltage Measurement... 08:37:20 Voltage Offset Correction: v_off_ini = -162.768 uV, v_off_fin = -160.275 uV 08:37:20 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -30.599 uVs 08:37:20 Integrated Voltage Measurement... 08:37:29 Voltage Offset Correction: v_off_ini = -147.862 uV, v_off_fin = -159.903 uV 08:37:29 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 21.120 uVs 08:37:29 Integrated Voltage Measurement... 08:37:38 Voltage Offset Correction: v_off_ini = -157.190 uV, v_off_fin = -155.990 uV 08:37:38 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -30.589 uVs 08:37:38 Integrated Voltage Measurement... 08:37:47 Voltage Offset Correction: v_off_ini = -153.316 uV, v_off_fin = -164.363 uV 08:37:47 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 3.777 uVs 08:37:47 Integrated Voltage Measurement... 08:37:56 Voltage Offset Correction: v_off_ini = -162.317 uV, v_off_fin = -152.428 uV 08:37:56 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -34.355 uVs 08:37:56 Integrated Voltage Measurement... 08:38:05 Voltage Offset Correction: v_off_ini = -158.510 uV, v_off_fin = -181.451 uV 08:38:05 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = 2.981 uVs 08:38:09 I1X Average and RMS of Measurements VT = -20.570 +- 3.789 uVs I1X = 4.114 +- 0.758 uTm 08:38:12 Wire Position: X0 = -2.0000 mm, Y0 = 0.0000 mm 08:38:16 Integrated Voltage Measurement... 08:38:25 Voltage Offset Correction: v_off_ini = -171.845 uV, v_off_fin = -197.119 uV 08:38:25 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -207.353 uVs 08:38:25 Integrated Voltage Measurement... 08:38:34 Voltage Offset Correction: v_off_ini = -175.259 uV, v_off_fin = -163.322 uV 08:38:34 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 219.721 uVs 08:38:34 Integrated Voltage Measurement... 08:38:43 Voltage Offset Correction: v_off_ini = -170.532 uV, v_off_fin = -178.625 uV 08:38:43 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -220.385 uVs 08:38:43 Integrated Voltage Measurement... 08:38:52 Voltage Offset Correction: v_off_ini = -170.391 uV, v_off_fin = -161.360 uV 08:38:52 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 215.059 uVs 08:38:52 Integrated Voltage Measurement... 08:39:01 Voltage Offset Correction: v_off_ini = -180.727 uV, v_off_fin = -176.160 uV 08:39:01 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -218.230 uVs 08:39:01 Integrated Voltage Measurement... 08:39:10 Voltage Offset Correction: v_off_ini = -166.702 uV, v_off_fin = -159.826 uV 08:39:10 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 231.465 uVs 08:39:14 I1Y Average and RMS of Measurements VT = -218.702 +- 4.669 uVs I1Y = -43.740 +- 0.934 uTm 08:39:19 Wire Position: X0 = -2.0000 mm, Y0 = 0.0000 mm 08:39:22 Integrated Voltage Measurement... 08:39:31 Voltage Offset Correction: v_off_ini = -166.525 uV, v_off_fin = -173.870 uV 08:39:31 Integrated Voltage Measurement dXYus = -1.0002 mm, dXYds = 0.0000 mm dXY = -1.0002 mm VT = -52.612 uVs 08:39:31 Integrated Voltage Measurement... 08:39:40 Voltage Offset Correction: v_off_ini = -171.894 uV, v_off_fin = -164.105 uV 08:39:40 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 44.009 uVs 08:39:40 Integrated Voltage Measurement... 08:39:49 Voltage Offset Correction: v_off_ini = -158.769 uV, v_off_fin = -152.284 uV 08:39:49 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -43.871 uVs 08:39:49 Integrated Voltage Measurement... 08:39:58 Voltage Offset Correction: v_off_ini = -163.976 uV, v_off_fin = -178.573 uV 08:39:58 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 24.093 uVs 08:39:58 Integrated Voltage Measurement... 08:40:07 Voltage Offset Correction: v_off_ini = -177.051 uV, v_off_fin = -159.731 uV 08:40:07 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -25.836 uVs 08:40:07 Integrated Voltage Measurement... 08:40:16 Voltage Offset Correction: v_off_ini = -175.540 uV, v_off_fin = -178.723 uV 08:40:16 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 30.297 uVs 08:40:18 I2X Average and RMS of Measurements VT = -36.786 +- 8.499 uVs I2X = 3.384 +- 0.782 uTm^2 08:40:24 Wire Position: X0 = -2.0000 mm, Y0 = 0.0000 mm 08:40:26 Integrated Voltage Measurement... 08:40:35 Voltage Offset Correction: v_off_ini = -174.937 uV, v_off_fin = -159.437 uV 08:40:35 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -211.026 uVs 08:40:35 Integrated Voltage Measurement... 08:40:44 Voltage Offset Correction: v_off_ini = -169.802 uV, v_off_fin = -156.474 uV 08:40:44 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 206.981 uVs 08:40:44 Integrated Voltage Measurement... 08:40:53 Voltage Offset Correction: v_off_ini = -181.053 uV, v_off_fin = -173.725 uV 08:40:53 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -202.585 uVs 08:40:53 Integrated Voltage Measurement... 08:41:02 Voltage Offset Correction: v_off_ini = -163.514 uV, v_off_fin = -162.016 uV 08:41:02 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 208.304 uVs 08:41:02 Integrated Voltage Measurement... 08:41:11 Voltage Offset Correction: v_off_ini = -178.165 uV, v_off_fin = -171.428 uV 08:41:11 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -201.933 uVs 08:41:11 Integrated Voltage Measurement... 08:41:20 Voltage Offset Correction: v_off_ini = -145.262 uV, v_off_fin = -146.927 uV 08:41:20 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 196.644 uVs 08:41:23 I2Y Average and RMS of Measurements VT = -204.579 +- 4.013 uVs I2Y = -18.821 +- 0.369 uTm^2 08:41:27 Wire Position: X0 = -1.0000 mm, Y0 = 0.0000 mm 08:41:32 Integrated Voltage Measurement... 08:41:40 Voltage Offset Correction: v_off_ini = -163.685 uV, v_off_fin = -161.766 uV 08:41:41 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -41.281 uVs 08:41:41 Integrated Voltage Measurement... 08:41:50 Voltage Offset Correction: v_off_ini = -167.785 uV, v_off_fin = -154.726 uV 08:41:50 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 65.609 uVs 08:41:50 Integrated Voltage Measurement... 08:41:59 Voltage Offset Correction: v_off_ini = -168.188 uV, v_off_fin = -156.788 uV 08:41:59 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -58.247 uVs 08:41:59 Integrated Voltage Measurement... 08:42:08 Voltage Offset Correction: v_off_ini = -163.929 uV, v_off_fin = -155.894 uV 08:42:08 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 57.018 uVs 08:42:08 Integrated Voltage Measurement... 08:42:17 Voltage Offset Correction: v_off_ini = -164.504 uV, v_off_fin = -167.902 uV 08:42:17 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -73.760 uVs 08:42:17 Integrated Voltage Measurement... 08:42:26 Voltage Offset Correction: v_off_ini = -162.233 uV, v_off_fin = -157.846 uV 08:42:26 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 63.220 uVs 08:42:30 I1X Average and RMS of Measurements VT = -59.856 +- 6.340 uVs I1X = 11.971 +- 1.268 uTm 08:42:32 Wire Position: X0 = -1.0000 mm, Y0 = 0.0000 mm 08:42:37 Integrated Voltage Measurement... 08:42:46 Voltage Offset Correction: v_off_ini = -144.800 uV, v_off_fin = -157.214 uV 08:42:46 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = -227.813 uVs 08:42:46 Integrated Voltage Measurement... 08:42:54 Voltage Offset Correction: v_off_ini = -173.238 uV, v_off_fin = -159.668 uV 08:42:55 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 231.314 uVs 08:42:55 Integrated Voltage Measurement... 08:43:03 Voltage Offset Correction: v_off_ini = -151.006 uV, v_off_fin = -167.110 uV 08:43:04 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.4999 mm dXY = 0.4999 mm VT = -221.657 uVs 08:43:04 Integrated Voltage Measurement... 08:43:12 Voltage Offset Correction: v_off_ini = -153.781 uV, v_off_fin = -149.601 uV 08:43:12 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 217.325 uVs 08:43:13 Integrated Voltage Measurement... 08:43:21 Voltage Offset Correction: v_off_ini = -172.627 uV, v_off_fin = -173.431 uV 08:43:22 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -215.306 uVs 08:43:22 Integrated Voltage Measurement... 08:43:30 Voltage Offset Correction: v_off_ini = -162.938 uV, v_off_fin = -141.501 uV 08:43:30 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 228.753 uVs 08:43:35 I1Y Average and RMS of Measurements VT = -223.695 +- 4.277 uVs I1Y = -44.741 +- 0.855 uTm 08:43:40 Wire Position: X0 = -1.0000 mm, Y0 = 0.0000 mm 08:43:42 Integrated Voltage Measurement... 08:43:51 Voltage Offset Correction: v_off_ini = -150.268 uV, v_off_fin = -148.591 uV 08:43:51 Integrated Voltage Measurement dXYus = -1.0002 mm, dXYds = 0.0000 mm dXY = -1.0002 mm VT = -45.916 uVs 08:43:51 Integrated Voltage Measurement... 08:44:00 Voltage Offset Correction: v_off_ini = -152.482 uV, v_off_fin = -150.581 uV 08:44:00 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 62.508 uVs 08:44:00 Integrated Voltage Measurement... 08:44:09 Voltage Offset Correction: v_off_ini = -164.932 uV, v_off_fin = -149.260 uV 08:44:09 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -57.914 uVs 08:44:10 Integrated Voltage Measurement... 08:44:18 Voltage Offset Correction: v_off_ini = -153.000 uV, v_off_fin = -161.436 uV 08:44:19 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 50.779 uVs 08:44:19 Integrated Voltage Measurement... 08:44:28 Voltage Offset Correction: v_off_ini = -136.606 uV, v_off_fin = -136.328 uV 08:44:28 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -62.983 uVs 08:44:28 Integrated Voltage Measurement... 08:44:37 Voltage Offset Correction: v_off_ini = -129.405 uV, v_off_fin = -144.577 uV 08:44:37 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 44.569 uVs 08:44:39 I2X Average and RMS of Measurements VT = -54.112 +- 0.244 uVs I2X = 4.978 +- 0.022 uTm^2 08:44:44 Wire Position: X0 = -1.0000 mm, Y0 = 0.0000 mm 08:44:47 Integrated Voltage Measurement... 08:44:56 Voltage Offset Correction: v_off_ini = -169.421 uV, v_off_fin = -169.861 uV 08:44:56 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -206.287 uVs 08:44:56 Integrated Voltage Measurement... 08:45:05 Voltage Offset Correction: v_off_ini = -144.593 uV, v_off_fin = -154.021 uV 08:45:05 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 209.002 uVs 08:45:05 Integrated Voltage Measurement... 08:45:14 Voltage Offset Correction: v_off_ini = -153.136 uV, v_off_fin = -151.997 uV 08:45:14 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -245.251 uVs 08:45:14 Integrated Voltage Measurement... 08:45:23 Voltage Offset Correction: v_off_ini = -149.145 uV, v_off_fin = -143.974 uV 08:45:23 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 216.691 uVs 08:45:23 Integrated Voltage Measurement... 08:45:32 Voltage Offset Correction: v_off_ini = -146.364 uV, v_off_fin = -167.013 uV 08:45:32 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -228.437 uVs 08:45:32 Integrated Voltage Measurement... 08:45:41 Voltage Offset Correction: v_off_ini = -176.308 uV, v_off_fin = -163.078 uV 08:45:41 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 212.821 uVs 08:45:43 I2Y Average and RMS of Measurements VT = -219.748 +- 9.543 uVs I2Y = -20.217 +- 0.878 uTm^2 08:45:48 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 08:45:52 Integrated Voltage Measurement... 08:46:01 Voltage Offset Correction: v_off_ini = -187.718 uV, v_off_fin = -196.353 uV 08:46:01 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -82.999 uVs 08:46:01 Integrated Voltage Measurement... 08:46:10 Voltage Offset Correction: v_off_ini = -166.122 uV, v_off_fin = -163.345 uV 08:46:10 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 66.011 uVs 08:46:10 Integrated Voltage Measurement... 08:46:19 Voltage Offset Correction: v_off_ini = -168.083 uV, v_off_fin = -178.400 uV 08:46:19 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -85.518 uVs 08:46:19 Integrated Voltage Measurement... 08:46:28 Voltage Offset Correction: v_off_ini = -163.153 uV, v_off_fin = -174.297 uV 08:46:28 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 66.542 uVs 08:46:28 Integrated Voltage Measurement... 08:46:37 Voltage Offset Correction: v_off_ini = -169.445 uV, v_off_fin = -172.625 uV 08:46:37 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -74.831 uVs 08:46:37 Integrated Voltage Measurement... 08:46:46 Voltage Offset Correction: v_off_ini = -183.281 uV, v_off_fin = -178.830 uV 08:46:46 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 73.140 uVs 08:46:50 I1X Average and RMS of Measurements VT = -74.840 +- 0.868 uVs I1X = 14.968 +- 0.174 uTm 08:46:53 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 08:46:57 Integrated Voltage Measurement... 08:47:06 Voltage Offset Correction: v_off_ini = -181.267 uV, v_off_fin = -187.285 uV 08:47:06 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -233.158 uVs 08:47:06 Integrated Voltage Measurement... 08:47:15 Voltage Offset Correction: v_off_ini = -174.796 uV, v_off_fin = -172.418 uV 08:47:15 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 240.201 uVs 08:47:15 Integrated Voltage Measurement... 08:47:24 Voltage Offset Correction: v_off_ini = -180.632 uV, v_off_fin = -188.445 uV 08:47:24 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -245.287 uVs 08:47:24 Integrated Voltage Measurement... 08:47:33 Voltage Offset Correction: v_off_ini = -198.628 uV, v_off_fin = -193.253 uV 08:47:33 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 252.025 uVs 08:47:33 Integrated Voltage Measurement... 08:47:42 Voltage Offset Correction: v_off_ini = -193.164 uV, v_off_fin = -194.989 uV 08:47:42 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -242.237 uVs 08:47:42 Integrated Voltage Measurement... 08:47:51 Voltage Offset Correction: v_off_ini = -177.282 uV, v_off_fin = -167.239 uV 08:47:51 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 249.181 uVs 08:47:55 I1Y Average and RMS of Measurements VT = -243.682 +- 5.095 uVs I1Y = -48.736 +- 1.019 uTm 08:48:00 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 08:48:03 Integrated Voltage Measurement... 08:48:12 Voltage Offset Correction: v_off_ini = -165.504 uV, v_off_fin = -177.891 uV 08:48:12 Integrated Voltage Measurement dXYus = -1.0002 mm, dXYds = 0.0000 mm dXY = -1.0002 mm VT = -112.217 uVs 08:48:12 Integrated Voltage Measurement... 08:48:21 Voltage Offset Correction: v_off_ini = -184.905 uV, v_off_fin = -197.120 uV 08:48:21 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 104.517 uVs 08:48:21 Integrated Voltage Measurement... 08:48:30 Voltage Offset Correction: v_off_ini = -181.442 uV, v_off_fin = -179.396 uV 08:48:30 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -90.856 uVs 08:48:30 Integrated Voltage Measurement... 08:48:39 Voltage Offset Correction: v_off_ini = -189.400 uV, v_off_fin = -182.132 uV 08:48:39 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 93.669 uVs 08:48:39 Integrated Voltage Measurement... 08:48:48 Voltage Offset Correction: v_off_ini = -190.903 uV, v_off_fin = -180.604 uV 08:48:48 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -87.368 uVs 08:48:48 Integrated Voltage Measurement... 08:48:57 Voltage Offset Correction: v_off_ini = -182.095 uV, v_off_fin = -191.241 uV 08:48:57 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 71.750 uVs 08:49:00 I2X Average and RMS of Measurements VT = -93.396 +- 11.788 uVs I2X = 8.592 +- 1.084 uTm^2 08:49:05 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 08:49:07 Integrated Voltage Measurement... 08:49:16 Voltage Offset Correction: v_off_ini = -188.606 uV, v_off_fin = -180.803 uV 08:49:16 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -233.544 uVs 08:49:16 Integrated Voltage Measurement... 08:49:25 Voltage Offset Correction: v_off_ini = -197.030 uV, v_off_fin = -179.307 uV 08:49:25 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 244.979 uVs 08:49:25 Integrated Voltage Measurement... 08:49:34 Voltage Offset Correction: v_off_ini = -198.956 uV, v_off_fin = -188.505 uV 08:49:34 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -222.434 uVs 08:49:34 Integrated Voltage Measurement... 08:49:43 Voltage Offset Correction: v_off_ini = -199.062 uV, v_off_fin = -180.626 uV 08:49:43 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 228.193 uVs 08:49:43 Integrated Voltage Measurement... 08:49:52 Voltage Offset Correction: v_off_ini = -174.031 uV, v_off_fin = -177.660 uV 08:49:52 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -240.377 uVs 08:49:52 Integrated Voltage Measurement... 08:50:01 Voltage Offset Correction: v_off_ini = -171.152 uV, v_off_fin = -183.320 uV 08:50:01 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 237.143 uVs 08:50:04 I2Y Average and RMS of Measurements VT = -234.445 +- 6.460 uVs I2Y = -21.569 +- 0.594 uTm^2 08:50:08 Wire Position: X0 = 1.0000 mm, Y0 = 0.0000 mm 08:50:13 Integrated Voltage Measurement... 08:50:22 Voltage Offset Correction: v_off_ini = -181.023 uV, v_off_fin = -180.057 uV 08:50:22 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -72.802 uVs 08:50:22 Integrated Voltage Measurement... 08:50:31 Voltage Offset Correction: v_off_ini = -184.791 uV, v_off_fin = -175.206 uV 08:50:31 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 115.098 uVs 08:50:31 Integrated Voltage Measurement... 08:50:40 Voltage Offset Correction: v_off_ini = -171.798 uV, v_off_fin = -183.385 uV 08:50:40 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -99.707 uVs 08:50:40 Integrated Voltage Measurement... 08:50:49 Voltage Offset Correction: v_off_ini = -164.391 uV, v_off_fin = -168.095 uV 08:50:49 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 97.936 uVs 08:50:49 Integrated Voltage Measurement... 08:50:58 Voltage Offset Correction: v_off_ini = -170.226 uV, v_off_fin = -176.847 uV 08:50:58 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -99.641 uVs 08:50:58 Integrated Voltage Measurement... 08:51:07 Voltage Offset Correction: v_off_ini = -166.281 uV, v_off_fin = -167.334 uV 08:51:07 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 100.189 uVs 08:51:11 I1X Average and RMS of Measurements VT = -97.562 +- 2.593 uVs I1X = 19.512 +- 0.519 uTm 08:51:13 Wire Position: X0 = 1.0000 mm, Y0 = 0.0000 mm 08:51:18 Integrated Voltage Measurement... 08:51:26 Voltage Offset Correction: v_off_ini = -173.849 uV, v_off_fin = -173.815 uV 08:51:26 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -221.241 uVs 08:51:27 Integrated Voltage Measurement... 08:51:35 Voltage Offset Correction: v_off_ini = -178.413 uV, v_off_fin = -167.394 uV 08:51:35 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 231.456 uVs 08:51:36 Integrated Voltage Measurement... 08:51:44 Voltage Offset Correction: v_off_ini = -166.118 uV, v_off_fin = -184.303 uV 08:51:44 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -235.230 uVs 08:51:44 Integrated Voltage Measurement... 08:51:53 Voltage Offset Correction: v_off_ini = -180.354 uV, v_off_fin = -172.112 uV 08:51:53 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 250.408 uVs 08:51:53 Integrated Voltage Measurement... 08:52:02 Voltage Offset Correction: v_off_ini = -177.347 uV, v_off_fin = -167.664 uV 08:52:02 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -244.524 uVs 08:52:02 Integrated Voltage Measurement... 08:52:11 Voltage Offset Correction: v_off_ini = -162.750 uV, v_off_fin = -152.310 uV 08:52:11 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 235.543 uVs 08:52:16 I1Y Average and RMS of Measurements VT = -236.401 +- 7.198 uVs I1Y = -47.280 +- 1.440 uTm 08:52:21 Wire Position: X0 = 1.0000 mm, Y0 = 0.0000 mm 08:52:23 Integrated Voltage Measurement... 08:52:32 Voltage Offset Correction: v_off_ini = -160.974 uV, v_off_fin = -153.615 uV 08:52:32 Integrated Voltage Measurement dXYus = -1.0002 mm, dXYds = 0.0000 mm dXY = -1.0002 mm VT = -101.228 uVs 08:52:32 Integrated Voltage Measurement... 08:52:41 Voltage Offset Correction: v_off_ini = -162.448 uV, v_off_fin = -165.444 uV 08:52:41 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 111.664 uVs 08:52:41 Integrated Voltage Measurement... 08:52:50 Voltage Offset Correction: v_off_ini = -154.741 uV, v_off_fin = -157.354 uV 08:52:50 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -103.101 uVs 08:52:50 Integrated Voltage Measurement... 08:52:59 Voltage Offset Correction: v_off_ini = -164.538 uV, v_off_fin = -176.920 uV 08:52:59 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 104.874 uVs 08:52:59 Integrated Voltage Measurement... 08:53:08 Voltage Offset Correction: v_off_ini = -151.848 uV, v_off_fin = -148.319 uV 08:53:08 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -109.606 uVs 08:53:08 Integrated Voltage Measurement... 08:53:17 Voltage Offset Correction: v_off_ini = -163.901 uV, v_off_fin = -168.011 uV 08:53:17 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 120.692 uVs 08:53:20 I2X Average and RMS of Measurements VT = -108.528 +- 4.789 uVs I2X = 9.984 +- 0.441 uTm^2 08:53:25 Wire Position: X0 = 1.0000 mm, Y0 = 0.0000 mm 08:53:28 Integrated Voltage Measurement... 08:53:36 Voltage Offset Correction: v_off_ini = -175.144 uV, v_off_fin = -174.846 uV 08:53:37 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -225.206 uVs 08:53:37 Integrated Voltage Measurement... 08:53:45 Voltage Offset Correction: v_off_ini = -174.849 uV, v_off_fin = -179.153 uV 08:53:46 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 228.076 uVs 08:53:46 Integrated Voltage Measurement... 08:53:55 Voltage Offset Correction: v_off_ini = -164.866 uV, v_off_fin = -173.063 uV 08:53:55 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -238.089 uVs 08:53:55 Integrated Voltage Measurement... 08:54:04 Voltage Offset Correction: v_off_ini = -162.161 uV, v_off_fin = -177.356 uV 08:54:04 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 230.086 uVs 08:54:04 Integrated Voltage Measurement... 08:54:13 Voltage Offset Correction: v_off_ini = -167.818 uV, v_off_fin = -187.291 uV 08:54:13 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -243.899 uVs 08:54:13 Integrated Voltage Measurement... 08:54:22 Voltage Offset Correction: v_off_ini = -174.435 uV, v_off_fin = -167.559 uV 08:54:22 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 220.174 uVs 08:54:24 I2Y Average and RMS of Measurements VT = -230.922 +- 3.140 uVs I2Y = -21.245 +- 0.289 uTm^2 08:54:29 Wire Position: X0 = 2.0000 mm, Y0 = 0.0000 mm 08:54:33 Integrated Voltage Measurement... 08:54:42 Voltage Offset Correction: v_off_ini = -169.589 uV, v_off_fin = -166.567 uV 08:54:42 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -116.826 uVs 08:54:42 Integrated Voltage Measurement... 08:54:51 Voltage Offset Correction: v_off_ini = -166.084 uV, v_off_fin = -159.431 uV 08:54:51 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 105.160 uVs 08:54:51 Integrated Voltage Measurement... 08:55:00 Voltage Offset Correction: v_off_ini = -153.898 uV, v_off_fin = -159.679 uV 08:55:00 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -120.301 uVs 08:55:00 Integrated Voltage Measurement... 08:55:09 Voltage Offset Correction: v_off_ini = -162.700 uV, v_off_fin = -139.285 uV 08:55:09 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 114.102 uVs 08:55:09 Integrated Voltage Measurement... 08:55:18 Voltage Offset Correction: v_off_ini = -137.228 uV, v_off_fin = -158.101 uV 08:55:18 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -88.980 uVs 08:55:18 Integrated Voltage Measurement... 08:55:27 Voltage Offset Correction: v_off_ini = -163.972 uV, v_off_fin = -154.727 uV 08:55:27 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 115.100 uVs 08:55:31 I1X Average and RMS of Measurements VT = -110.078 +- 6.223 uVs I1X = 22.016 +- 1.245 uTm 08:55:34 Wire Position: X0 = 2.0000 mm, Y0 = 0.0000 mm 08:55:38 Integrated Voltage Measurement... 08:55:47 Voltage Offset Correction: v_off_ini = -142.420 uV, v_off_fin = -152.807 uV 08:55:47 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -251.046 uVs 08:55:47 Integrated Voltage Measurement... 08:55:56 Voltage Offset Correction: v_off_ini = -131.302 uV, v_off_fin = -126.580 uV 08:55:56 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 212.512 uVs 08:55:56 Integrated Voltage Measurement... 08:56:05 Voltage Offset Correction: v_off_ini = -142.681 uV, v_off_fin = -143.337 uV 08:56:05 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -241.804 uVs 08:56:05 Integrated Voltage Measurement... 08:56:14 Voltage Offset Correction: v_off_ini = -139.456 uV, v_off_fin = -138.704 uV 08:56:14 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.4999 mm dXY = -0.4999 mm VT = 244.580 uVs 08:56:14 Integrated Voltage Measurement... 08:56:23 Voltage Offset Correction: v_off_ini = -148.817 uV, v_off_fin = -140.123 uV 08:56:23 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -249.319 uVs 08:56:23 Integrated Voltage Measurement... 08:56:32 Voltage Offset Correction: v_off_ini = -142.579 uV, v_off_fin = -129.800 uV 08:56:32 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 244.835 uVs 08:56:36 I1Y Average and RMS of Measurements VT = -240.683 +- 6.493 uVs I1Y = -48.138 +- 1.299 uTm 08:56:41 Wire Position: X0 = 2.0000 mm, Y0 = 0.0000 mm 08:56:44 Integrated Voltage Measurement... 08:56:53 Voltage Offset Correction: v_off_ini = -123.971 uV, v_off_fin = -139.268 uV 08:56:53 Integrated Voltage Measurement dXYus = -1.0002 mm, dXYds = 0.0000 mm dXY = -1.0002 mm VT = -140.779 uVs 08:56:53 Integrated Voltage Measurement... 08:57:02 Voltage Offset Correction: v_off_ini = -134.955 uV, v_off_fin = -149.625 uV 08:57:02 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 134.420 uVs 08:57:02 Integrated Voltage Measurement... 08:57:11 Voltage Offset Correction: v_off_ini = -147.898 uV, v_off_fin = -126.774 uV 08:57:11 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -116.673 uVs 08:57:11 Integrated Voltage Measurement... 08:57:20 Voltage Offset Correction: v_off_ini = -139.103 uV, v_off_fin = -139.944 uV 08:57:20 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 132.723 uVs 08:57:20 Integrated Voltage Measurement... 08:57:29 Voltage Offset Correction: v_off_ini = -135.896 uV, v_off_fin = -137.536 uV 08:57:29 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -126.109 uVs 08:57:29 Integrated Voltage Measurement... 08:57:38 Voltage Offset Correction: v_off_ini = -141.989 uV, v_off_fin = -150.776 uV 08:57:38 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 129.801 uVs 08:57:40 I2X Average and RMS of Measurements VT = -130.084 +- 5.478 uVs I2X = 11.967 +- 0.503 uTm^2 08:57:45 Wire Position: X0 = 2.0000 mm, Y0 = 0.0000 mm 08:57:48 Integrated Voltage Measurement... 08:57:57 Voltage Offset Correction: v_off_ini = -135.734 uV, v_off_fin = -141.896 uV 08:57:57 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -226.225 uVs 08:57:57 Integrated Voltage Measurement... 08:58:06 Voltage Offset Correction: v_off_ini = -160.919 uV, v_off_fin = -141.010 uV 08:58:06 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 257.563 uVs 08:58:06 Integrated Voltage Measurement... 08:58:15 Voltage Offset Correction: v_off_ini = -153.361 uV, v_off_fin = -153.056 uV 08:58:15 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -207.329 uVs 08:58:15 Integrated Voltage Measurement... 08:58:24 Voltage Offset Correction: v_off_ini = -125.385 uV, v_off_fin = -138.094 uV 08:58:24 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 225.168 uVs 08:58:24 Integrated Voltage Measurement... 08:58:33 Voltage Offset Correction: v_off_ini = -166.740 uV, v_off_fin = -153.145 uV 08:58:33 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -237.655 uVs 08:58:33 Integrated Voltage Measurement... 08:58:42 Voltage Offset Correction: v_off_ini = -157.464 uV, v_off_fin = -152.776 uV 08:58:42 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 245.283 uVs 08:58:44 I2Y Average and RMS of Measurements VT = -233.204 +- 11.991 uVs I2Y = -21.455 +- 1.103 uTm^2 08:58:52 Wire Position: X0 = 0.0000 mm, Y0 = -1.0000 mm 08:58:56 Integrated Voltage Measurement... 08:59:05 Voltage Offset Correction: v_off_ini = -165.351 uV, v_off_fin = -157.772 uV 08:59:05 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -90.053 uVs 08:59:05 Integrated Voltage Measurement... 08:59:14 Voltage Offset Correction: v_off_ini = -162.591 uV, v_off_fin = -150.786 uV 08:59:14 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 92.820 uVs 08:59:14 Integrated Voltage Measurement... 08:59:23 Voltage Offset Correction: v_off_ini = -150.821 uV, v_off_fin = -134.121 uV 08:59:23 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -98.836 uVs 08:59:23 Integrated Voltage Measurement... 08:59:32 Voltage Offset Correction: v_off_ini = -147.916 uV, v_off_fin = -149.628 uV 08:59:32 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 92.209 uVs 08:59:32 Integrated Voltage Measurement... 08:59:41 Voltage Offset Correction: v_off_ini = -135.388 uV, v_off_fin = -148.071 uV 08:59:41 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -80.212 uVs 08:59:41 Integrated Voltage Measurement... 08:59:50 Voltage Offset Correction: v_off_ini = -143.767 uV, v_off_fin = -137.473 uV 08:59:50 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 96.781 uVs 08:59:54 I1X Average and RMS of Measurements VT = -91.819 +- 2.881 uVs I1X = 18.364 +- 0.576 uTm 08:59:57 Wire Position: X0 = 0.0000 mm, Y0 = -1.0000 mm 09:00:01 Integrated Voltage Measurement... 09:00:10 Voltage Offset Correction: v_off_ini = -144.087 uV, v_off_fin = -170.728 uV 09:00:10 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -230.395 uVs 09:00:10 Integrated Voltage Measurement... 09:00:19 Voltage Offset Correction: v_off_ini = -156.654 uV, v_off_fin = -139.542 uV 09:00:19 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 226.163 uVs 09:00:19 Integrated Voltage Measurement... 09:00:28 Voltage Offset Correction: v_off_ini = -137.604 uV, v_off_fin = -155.888 uV 09:00:28 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -216.893 uVs 09:00:28 Integrated Voltage Measurement... 09:00:37 Voltage Offset Correction: v_off_ini = -158.373 uV, v_off_fin = -145.913 uV 09:00:37 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 222.908 uVs 09:00:37 Integrated Voltage Measurement... 09:00:46 Voltage Offset Correction: v_off_ini = -155.875 uV, v_off_fin = -158.988 uV 09:00:46 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -220.959 uVs 09:00:46 Integrated Voltage Measurement... 09:00:55 Voltage Offset Correction: v_off_ini = -155.100 uV, v_off_fin = -136.377 uV 09:00:55 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 212.419 uVs 09:00:59 I1Y Average and RMS of Measurements VT = -221.623 +- 4.886 uVs I1Y = -44.325 +- 0.977 uTm 09:01:02 Wire Position: X0 = 0.0000 mm, Y0 = -1.0000 mm 09:01:04 Integrated Voltage Measurement... 09:01:13 Voltage Offset Correction: v_off_ini = -152.985 uV, v_off_fin = -145.264 uV 09:01:13 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -69.061 uVs 09:01:13 Integrated Voltage Measurement... 09:01:22 Voltage Offset Correction: v_off_ini = -150.599 uV, v_off_fin = -146.272 uV 09:01:22 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 72.197 uVs 09:01:22 Integrated Voltage Measurement... 09:01:31 Voltage Offset Correction: v_off_ini = -134.822 uV, v_off_fin = -140.557 uV 09:01:31 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -63.060 uVs 09:01:31 Integrated Voltage Measurement... 09:01:40 Voltage Offset Correction: v_off_ini = -131.786 uV, v_off_fin = -146.890 uV 09:01:40 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 93.940 uVs 09:01:40 Integrated Voltage Measurement... 09:01:49 Voltage Offset Correction: v_off_ini = -146.756 uV, v_off_fin = -135.620 uV 09:01:49 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -72.037 uVs 09:01:49 Integrated Voltage Measurement... 09:01:58 Voltage Offset Correction: v_off_ini = -147.662 uV, v_off_fin = -141.843 uV 09:01:58 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 78.237 uVs 09:02:01 I2X Average and RMS of Measurements VT = -74.755 +- 3.225 uVs I2X = 6.877 +- 0.297 uTm^2 09:02:03 Wire Position: X0 = 0.0000 mm, Y0 = -1.0000 mm 09:02:06 Integrated Voltage Measurement... 09:02:15 Voltage Offset Correction: v_off_ini = -135.124 uV, v_off_fin = -139.708 uV 09:02:15 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -202.255 uVs 09:02:15 Integrated Voltage Measurement... 09:02:24 Voltage Offset Correction: v_off_ini = -149.291 uV, v_off_fin = -141.132 uV 09:02:24 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 206.803 uVs 09:02:24 Integrated Voltage Measurement... 09:02:33 Voltage Offset Correction: v_off_ini = -142.390 uV, v_off_fin = -142.089 uV 09:02:33 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -216.566 uVs 09:02:33 Integrated Voltage Measurement... 09:02:42 Voltage Offset Correction: v_off_ini = -132.776 uV, v_off_fin = -136.998 uV 09:02:42 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 210.527 uVs 09:02:42 Integrated Voltage Measurement... 09:02:51 Voltage Offset Correction: v_off_ini = -126.681 uV, v_off_fin = -133.586 uV 09:02:51 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -230.215 uVs 09:02:51 Integrated Voltage Measurement... 09:03:00 Voltage Offset Correction: v_off_ini = -142.810 uV, v_off_fin = -145.716 uV 09:03:00 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 203.243 uVs 09:03:02 I2Y Average and RMS of Measurements VT = -211.601 +- 5.167 uVs I2Y = -19.467 +- 0.475 uTm^2 09:03:06 Wire Position: X0 = 0.0000 mm, Y0 = -0.5000 mm 09:03:11 Integrated Voltage Measurement... 09:03:20 Voltage Offset Correction: v_off_ini = -154.505 uV, v_off_fin = -145.034 uV 09:03:20 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -83.510 uVs 09:03:20 Integrated Voltage Measurement... 09:03:29 Voltage Offset Correction: v_off_ini = -149.883 uV, v_off_fin = -155.466 uV 09:03:29 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 79.191 uVs 09:03:29 Integrated Voltage Measurement... 09:03:38 Voltage Offset Correction: v_off_ini = -158.673 uV, v_off_fin = -156.918 uV 09:03:38 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -78.450 uVs 09:03:38 Integrated Voltage Measurement... 09:03:47 Voltage Offset Correction: v_off_ini = -152.289 uV, v_off_fin = -144.925 uV 09:03:47 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 89.470 uVs 09:03:47 Integrated Voltage Measurement... 09:03:56 Voltage Offset Correction: v_off_ini = -162.230 uV, v_off_fin = -153.136 uV 09:03:56 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = -84.381 uVs 09:03:56 Integrated Voltage Measurement... 09:04:05 Voltage Offset Correction: v_off_ini = -145.205 uV, v_off_fin = -157.842 uV 09:04:05 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = 88.104 uVs 09:04:09 I1X Average and RMS of Measurements VT = -83.851 +- 1.999 uVs I1X = 16.770 +- 0.399 uTm 09:04:11 Wire Position: X0 = 0.0000 mm, Y0 = -0.5000 mm 09:04:16 Integrated Voltage Measurement... 09:04:25 Voltage Offset Correction: v_off_ini = -135.988 uV, v_off_fin = -150.329 uV 09:04:25 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -231.810 uVs 09:04:25 Integrated Voltage Measurement... 09:04:33 Voltage Offset Correction: v_off_ini = -132.533 uV, v_off_fin = -116.146 uV 09:04:34 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.4999 mm dXY = -0.4999 mm VT = 229.507 uVs 09:04:34 Integrated Voltage Measurement... 09:04:42 Voltage Offset Correction: v_off_ini = -146.833 uV, v_off_fin = -155.270 uV 09:04:43 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -222.881 uVs 09:04:43 Integrated Voltage Measurement... 09:04:51 Voltage Offset Correction: v_off_ini = -143.375 uV, v_off_fin = -125.910 uV 09:04:51 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 241.277 uVs 09:04:52 Integrated Voltage Measurement... 09:05:00 Voltage Offset Correction: v_off_ini = -129.446 uV, v_off_fin = -142.331 uV 09:05:00 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -237.602 uVs 09:05:00 Integrated Voltage Measurement... 09:05:09 Voltage Offset Correction: v_off_ini = -120.177 uV, v_off_fin = -118.575 uV 09:05:09 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 238.602 uVs 09:05:14 I1Y Average and RMS of Measurements VT = -233.613 +- 3.227 uVs I1Y = -46.725 +- 0.644 uTm 09:05:16 Wire Position: X0 = 0.0000 mm, Y0 = -0.5000 mm 09:05:19 Integrated Voltage Measurement... 09:05:28 Voltage Offset Correction: v_off_ini = -130.154 uV, v_off_fin = -122.345 uV 09:05:28 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -72.234 uVs 09:05:28 Integrated Voltage Measurement... 09:05:37 Voltage Offset Correction: v_off_ini = -129.063 uV, v_off_fin = -135.862 uV 09:05:37 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 81.721 uVs 09:05:37 Integrated Voltage Measurement... 09:05:46 Voltage Offset Correction: v_off_ini = -128.472 uV, v_off_fin = -119.174 uV 09:05:46 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -88.895 uVs 09:05:46 Integrated Voltage Measurement... 09:05:55 Voltage Offset Correction: v_off_ini = -121.389 uV, v_off_fin = -142.009 uV 09:05:55 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 75.657 uVs 09:05:55 Integrated Voltage Measurement... 09:06:04 Voltage Offset Correction: v_off_ini = -134.700 uV, v_off_fin = -125.821 uV 09:06:04 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -82.397 uVs 09:06:04 Integrated Voltage Measurement... 09:06:13 Voltage Offset Correction: v_off_ini = -118.060 uV, v_off_fin = -125.594 uV 09:06:13 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 70.175 uVs 09:06:15 I2X Average and RMS of Measurements VT = -78.513 +- 2.676 uVs I2X = 7.223 +- 0.246 uTm^2 09:06:18 Wire Position: X0 = 0.0000 mm, Y0 = -0.5000 mm 09:06:20 Integrated Voltage Measurement... 09:06:29 Voltage Offset Correction: v_off_ini = -128.789 uV, v_off_fin = -127.246 uV 09:06:29 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -219.834 uVs 09:06:29 Integrated Voltage Measurement... 09:06:38 Voltage Offset Correction: v_off_ini = -127.583 uV, v_off_fin = -139.170 uV 09:06:38 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 248.361 uVs 09:06:38 Integrated Voltage Measurement... 09:06:47 Voltage Offset Correction: v_off_ini = -144.184 uV, v_off_fin = -130.391 uV 09:06:47 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -219.493 uVs 09:06:47 Integrated Voltage Measurement... 09:06:56 Voltage Offset Correction: v_off_ini = -124.983 uV, v_off_fin = -134.599 uV 09:06:56 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 206.958 uVs 09:06:57 Integrated Voltage Measurement... 09:07:05 Voltage Offset Correction: v_off_ini = -143.250 uV, v_off_fin = -147.927 uV 09:07:06 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -221.692 uVs 09:07:06 Integrated Voltage Measurement... 09:07:15 Voltage Offset Correction: v_off_ini = -143.942 uV, v_off_fin = -140.445 uV 09:07:15 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 216.254 uVs 09:07:17 I2Y Average and RMS of Measurements VT = -222.099 +- 8.803 uVs I2Y = -20.433 +- 0.810 uTm^2 09:07:21 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:07:25 Integrated Voltage Measurement... 09:07:34 Voltage Offset Correction: v_off_ini = -136.441 uV, v_off_fin = -139.058 uV 09:07:34 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -85.987 uVs 09:07:34 Integrated Voltage Measurement... 09:07:43 Voltage Offset Correction: v_off_ini = -151.302 uV, v_off_fin = -136.775 uV 09:07:43 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 93.446 uVs 09:07:43 Integrated Voltage Measurement... 09:07:52 Voltage Offset Correction: v_off_ini = -135.796 uV, v_off_fin = -140.505 uV 09:07:52 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -72.641 uVs 09:07:52 Integrated Voltage Measurement... 09:08:01 Voltage Offset Correction: v_off_ini = -139.683 uV, v_off_fin = -138.435 uV 09:08:01 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 84.760 uVs 09:08:01 Integrated Voltage Measurement... 09:08:10 Voltage Offset Correction: v_off_ini = -130.034 uV, v_off_fin = -135.407 uV 09:08:10 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -84.974 uVs 09:08:10 Integrated Voltage Measurement... 09:08:19 Voltage Offset Correction: v_off_ini = -130.723 uV, v_off_fin = -137.279 uV 09:08:19 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 83.297 uVs 09:08:23 I1X Average and RMS of Measurements VT = -84.184 +- 4.497 uVs I1X = 16.837 +- 0.900 uTm 09:08:26 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:08:30 Integrated Voltage Measurement... 09:08:39 Voltage Offset Correction: v_off_ini = -125.335 uV, v_off_fin = -150.961 uV 09:08:39 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -242.073 uVs 09:08:39 Integrated Voltage Measurement... 09:08:48 Voltage Offset Correction: v_off_ini = -127.778 uV, v_off_fin = -125.882 uV 09:08:48 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 225.475 uVs 09:08:48 Integrated Voltage Measurement... 09:08:57 Voltage Offset Correction: v_off_ini = -121.966 uV, v_off_fin = -132.849 uV 09:08:57 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -248.753 uVs 09:08:57 Integrated Voltage Measurement... 09:09:06 Voltage Offset Correction: v_off_ini = -115.445 uV, v_off_fin = -115.989 uV 09:09:06 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 234.250 uVs 09:09:06 Integrated Voltage Measurement... 09:09:15 Voltage Offset Correction: v_off_ini = -128.330 uV, v_off_fin = -148.526 uV 09:09:15 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.4999 mm dXY = 0.4999 mm VT = -219.474 uVs 09:09:15 Integrated Voltage Measurement... 09:09:24 Voltage Offset Correction: v_off_ini = -125.542 uV, v_off_fin = -113.637 uV 09:09:24 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.4999 mm dXY = -0.4999 mm VT = 225.454 uVs 09:09:28 I1Y Average and RMS of Measurements VT = -232.580 +- 7.818 uVs I1Y = -46.519 +- 1.560 uTm 09:09:31 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:09:34 Integrated Voltage Measurement... 09:09:43 Voltage Offset Correction: v_off_ini = -107.999 uV, v_off_fin = -111.276 uV 09:09:43 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -89.040 uVs 09:09:43 Integrated Voltage Measurement... 09:09:52 Voltage Offset Correction: v_off_ini = -126.936 uV, v_off_fin = -119.205 uV 09:09:52 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 99.258 uVs 09:09:52 Integrated Voltage Measurement... 09:10:01 Voltage Offset Correction: v_off_ini = -111.506 uV, v_off_fin = -95.440 uV 09:10:01 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -82.702 uVs 09:10:01 Integrated Voltage Measurement... 09:10:10 Voltage Offset Correction: v_off_ini = -111.182 uV, v_off_fin = -101.347 uV 09:10:10 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 89.522 uVs 09:10:10 Integrated Voltage Measurement... 09:10:19 Voltage Offset Correction: v_off_ini = -115.424 uV, v_off_fin = -108.991 uV 09:10:19 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -80.174 uVs 09:10:19 Integrated Voltage Measurement... 09:10:28 Voltage Offset Correction: v_off_ini = -107.212 uV, v_off_fin = -117.456 uV 09:10:28 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 80.421 uVs 09:10:30 I2X Average and RMS of Measurements VT = -86.853 +- 5.679 uVs I2X = 7.990 +- 0.522 uTm^2 09:10:33 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:10:35 Integrated Voltage Measurement... 09:10:44 Voltage Offset Correction: v_off_ini = -130.029 uV, v_off_fin = -124.747 uV 09:10:44 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -236.572 uVs 09:10:44 Integrated Voltage Measurement... 09:10:53 Voltage Offset Correction: v_off_ini = -120.306 uV, v_off_fin = -122.149 uV 09:10:53 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 233.089 uVs 09:10:53 Integrated Voltage Measurement... 09:11:02 Voltage Offset Correction: v_off_ini = -125.639 uV, v_off_fin = -109.512 uV 09:11:02 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -232.131 uVs 09:11:02 Integrated Voltage Measurement... 09:11:11 Voltage Offset Correction: v_off_ini = -101.038 uV, v_off_fin = -105.680 uV 09:11:11 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 225.299 uVs 09:11:11 Integrated Voltage Measurement... 09:11:20 Voltage Offset Correction: v_off_ini = -111.246 uV, v_off_fin = -105.867 uV 09:11:21 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -231.942 uVs 09:11:21 Integrated Voltage Measurement... 09:11:30 Voltage Offset Correction: v_off_ini = -89.485 uV, v_off_fin = -99.762 uV 09:11:30 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 238.061 uVs 09:11:32 I2Y Average and RMS of Measurements VT = -232.849 +- 2.924 uVs I2Y = -21.422 +- 0.269 uTm^2 09:11:36 Wire Position: X0 = 0.0000 mm, Y0 = 0.5000 mm 09:11:40 Integrated Voltage Measurement... 09:11:49 Voltage Offset Correction: v_off_ini = -97.557 uV, v_off_fin = -103.075 uV 09:11:49 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -61.789 uVs 09:11:49 Integrated Voltage Measurement... 09:11:58 Voltage Offset Correction: v_off_ini = -88.637 uV, v_off_fin = -89.538 uV 09:11:58 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 71.302 uVs 09:11:58 Integrated Voltage Measurement... 09:12:07 Voltage Offset Correction: v_off_ini = -99.251 uV, v_off_fin = -107.341 uV 09:12:07 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -48.096 uVs 09:12:07 Integrated Voltage Measurement... 09:12:16 Voltage Offset Correction: v_off_ini = -92.873 uV, v_off_fin = -88.569 uV 09:12:16 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 61.312 uVs 09:12:16 Integrated Voltage Measurement... 09:12:25 Voltage Offset Correction: v_off_ini = -97.981 uV, v_off_fin = -85.253 uV 09:12:25 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -80.118 uVs 09:12:25 Integrated Voltage Measurement... 09:12:34 Voltage Offset Correction: v_off_ini = -92.535 uV, v_off_fin = -82.741 uV 09:12:34 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 59.210 uVs 09:12:39 I1X Average and RMS of Measurements VT = -63.638 +- 6.444 uVs I1X = 12.728 +- 1.289 uTm 09:12:41 Wire Position: X0 = 0.0000 mm, Y0 = 0.5000 mm 09:12:45 Integrated Voltage Measurement... 09:12:54 Voltage Offset Correction: v_off_ini = -94.847 uV, v_off_fin = -113.171 uV 09:12:54 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -260.007 uVs 09:12:54 Integrated Voltage Measurement... 09:13:03 Voltage Offset Correction: v_off_ini = -91.365 uV, v_off_fin = -78.813 uV 09:13:03 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 265.004 uVs 09:13:03 Integrated Voltage Measurement... 09:13:12 Voltage Offset Correction: v_off_ini = -94.248 uV, v_off_fin = -108.295 uV 09:13:12 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -249.875 uVs 09:13:12 Integrated Voltage Measurement... 09:13:21 Voltage Offset Correction: v_off_ini = -94.277 uV, v_off_fin = -77.952 uV 09:13:21 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 241.781 uVs 09:13:21 Integrated Voltage Measurement... 09:13:30 Voltage Offset Correction: v_off_ini = -89.602 uV, v_off_fin = -99.811 uV 09:13:30 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -235.559 uVs 09:13:30 Integrated Voltage Measurement... 09:13:39 Voltage Offset Correction: v_off_ini = -99.922 uV, v_off_fin = -93.710 uV 09:13:39 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 235.080 uVs 09:13:43 I1Y Average and RMS of Measurements VT = -247.884 +- 11.194 uVs I1Y = -49.577 +- 2.239 uTm 09:13:46 Wire Position: X0 = 0.0000 mm, Y0 = 0.5000 mm 09:13:49 Integrated Voltage Measurement... 09:13:58 Voltage Offset Correction: v_off_ini = -80.453 uV, v_off_fin = -90.887 uV 09:13:58 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -62.704 uVs 09:13:58 Integrated Voltage Measurement... 09:14:07 Voltage Offset Correction: v_off_ini = -98.615 uV, v_off_fin = -100.451 uV 09:14:07 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 83.049 uVs 09:14:07 Integrated Voltage Measurement... 09:14:16 Voltage Offset Correction: v_off_ini = -87.018 uV, v_off_fin = -78.342 uV 09:14:16 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -88.448 uVs 09:14:16 Integrated Voltage Measurement... 09:14:25 Voltage Offset Correction: v_off_ini = -97.306 uV, v_off_fin = -90.870 uV 09:14:25 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 78.616 uVs 09:14:25 Integrated Voltage Measurement... 09:14:34 Voltage Offset Correction: v_off_ini = -81.736 uV, v_off_fin = -70.916 uV 09:14:34 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -100.027 uVs 09:14:34 Integrated Voltage Measurement... 09:14:43 Voltage Offset Correction: v_off_ini = -85.719 uV, v_off_fin = -90.653 uV 09:14:43 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 77.004 uVs 09:14:45 I2X Average and RMS of Measurements VT = -81.641 +- 6.523 uVs I2X = 7.511 +- 0.600 uTm^2 09:14:48 Wire Position: X0 = 0.0000 mm, Y0 = 0.5000 mm 09:14:50 Integrated Voltage Measurement... 09:14:59 Voltage Offset Correction: v_off_ini = -76.001 uV, v_off_fin = -82.006 uV 09:14:59 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -262.950 uVs 09:14:59 Integrated Voltage Measurement... 09:15:08 Voltage Offset Correction: v_off_ini = -68.724 uV, v_off_fin = -77.793 uV 09:15:08 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 240.325 uVs 09:15:08 Integrated Voltage Measurement... 09:15:17 Voltage Offset Correction: v_off_ini = -77.204 uV, v_off_fin = -89.615 uV 09:15:17 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -229.535 uVs 09:15:17 Integrated Voltage Measurement... 09:15:26 Voltage Offset Correction: v_off_ini = -73.445 uV, v_off_fin = -73.939 uV 09:15:26 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 242.652 uVs 09:15:27 Integrated Voltage Measurement... 09:15:35 Voltage Offset Correction: v_off_ini = -71.000 uV, v_off_fin = -71.559 uV 09:15:36 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -239.433 uVs 09:15:36 Integrated Voltage Measurement... 09:15:45 Voltage Offset Correction: v_off_ini = -69.403 uV, v_off_fin = -95.735 uV 09:15:45 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 246.606 uVs 09:15:47 I2Y Average and RMS of Measurements VT = -243.583 +- 6.358 uVs I2Y = -22.410 +- 0.585 uTm^2 09:15:51 Wire Position: X0 = 0.0000 mm, Y0 = 1.0000 mm 09:15:55 Integrated Voltage Measurement... 09:16:04 Voltage Offset Correction: v_off_ini = -88.238 uV, v_off_fin = -89.366 uV 09:16:04 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -77.637 uVs 09:16:04 Integrated Voltage Measurement... 09:16:13 Voltage Offset Correction: v_off_ini = -84.986 uV, v_off_fin = -79.324 uV 09:16:13 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 39.527 uVs 09:16:13 Integrated Voltage Measurement... 09:16:22 Voltage Offset Correction: v_off_ini = -80.892 uV, v_off_fin = -92.493 uV 09:16:22 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -69.670 uVs 09:16:22 Integrated Voltage Measurement... 09:16:31 Voltage Offset Correction: v_off_ini = -101.080 uV, v_off_fin = -93.352 uV 09:16:31 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 83.708 uVs 09:16:31 Integrated Voltage Measurement... 09:16:40 Voltage Offset Correction: v_off_ini = -83.045 uV, v_off_fin = -79.418 uV 09:16:40 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -84.798 uVs 09:16:40 Integrated Voltage Measurement... 09:16:49 Voltage Offset Correction: v_off_ini = -85.224 uV, v_off_fin = -87.294 uV 09:16:49 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 82.850 uVs 09:16:54 I1X Average and RMS of Measurements VT = -73.032 +- 10.624 uVs I1X = 14.607 +- 2.125 uTm 09:16:56 Wire Position: X0 = 0.0000 mm, Y0 = 1.0000 mm 09:17:00 Integrated Voltage Measurement... 09:17:09 Voltage Offset Correction: v_off_ini = -104.675 uV, v_off_fin = -113.514 uV 09:17:09 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -254.819 uVs 09:17:09 Integrated Voltage Measurement... 09:17:18 Voltage Offset Correction: v_off_ini = -117.604 uV, v_off_fin = -108.903 uV 09:17:18 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 261.506 uVs 09:17:18 Integrated Voltage Measurement... 09:17:27 Voltage Offset Correction: v_off_ini = -103.596 uV, v_off_fin = -122.626 uV 09:17:27 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -273.803 uVs 09:17:27 Integrated Voltage Measurement... 09:17:36 Voltage Offset Correction: v_off_ini = -102.190 uV, v_off_fin = -93.460 uV 09:17:36 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 252.973 uVs 09:17:36 Integrated Voltage Measurement... 09:17:45 Voltage Offset Correction: v_off_ini = -104.526 uV, v_off_fin = -104.794 uV 09:17:45 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -258.965 uVs 09:17:45 Integrated Voltage Measurement... 09:17:54 Voltage Offset Correction: v_off_ini = -107.175 uV, v_off_fin = -81.534 uV 09:17:54 Integrated Voltage Measurement dXYus = -0.5001 mm, dXYds = -0.5001 mm dXY = -0.5001 mm VT = 250.418 uVs 09:17:59 I1Y Average and RMS of Measurements VT = -258.747 +- 3.574 uVs I1Y = -51.748 +- 0.717 uTm 09:18:01 Wire Position: X0 = 0.0000 mm, Y0 = 1.0000 mm 09:18:04 Integrated Voltage Measurement... 09:18:13 Voltage Offset Correction: v_off_ini = -96.588 uV, v_off_fin = -115.440 uV 09:18:13 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -51.453 uVs 09:18:13 Integrated Voltage Measurement... 09:18:22 Voltage Offset Correction: v_off_ini = -105.982 uV, v_off_fin = -124.441 uV 09:18:22 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 86.606 uVs 09:18:22 Integrated Voltage Measurement... 09:18:31 Voltage Offset Correction: v_off_ini = -116.031 uV, v_off_fin = -115.129 uV 09:18:31 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -69.814 uVs 09:18:31 Integrated Voltage Measurement... 09:18:40 Voltage Offset Correction: v_off_ini = -110.113 uV, v_off_fin = -119.580 uV 09:18:40 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 63.624 uVs 09:18:40 Integrated Voltage Measurement... 09:18:49 Voltage Offset Correction: v_off_ini = -126.494 uV, v_off_fin = -118.468 uV 09:18:49 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -58.489 uVs 09:18:49 Integrated Voltage Measurement... 09:18:58 Voltage Offset Correction: v_off_ini = -113.155 uV, v_off_fin = -116.975 uV 09:18:58 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 64.686 uVs 09:19:00 I2X Average and RMS of Measurements VT = -65.779 +- 3.110 uVs I2X = 6.052 +- 0.286 uTm^2 09:19:03 Wire Position: X0 = 0.0000 mm, Y0 = 1.0000 mm 09:19:05 Integrated Voltage Measurement... 09:19:14 Voltage Offset Correction: v_off_ini = -117.158 uV, v_off_fin = -112.409 uV 09:19:14 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -250.346 uVs 09:19:14 Integrated Voltage Measurement... 09:19:23 Voltage Offset Correction: v_off_ini = -126.370 uV, v_off_fin = -102.143 uV 09:19:23 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 256.331 uVs 09:19:23 Integrated Voltage Measurement... 09:19:32 Voltage Offset Correction: v_off_ini = -107.058 uV, v_off_fin = -102.110 uV 09:19:33 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -225.345 uVs 09:19:33 Integrated Voltage Measurement... 09:19:42 Voltage Offset Correction: v_off_ini = -105.870 uV, v_off_fin = -98.023 uV 09:19:42 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 248.857 uVs 09:19:42 Integrated Voltage Measurement... 09:19:51 Voltage Offset Correction: v_off_ini = -102.528 uV, v_off_fin = -90.487 uV 09:19:51 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -249.313 uVs 09:19:51 Integrated Voltage Measurement... 09:20:00 Voltage Offset Correction: v_off_ini = -102.658 uV, v_off_fin = -97.523 uV 09:20:00 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 250.399 uVs 09:20:02 I2Y Average and RMS of Measurements VT = -246.765 +- 6.980 uVs I2Y = -22.702 +- 0.642 uTm^2 09:20:17 Wire Position: X0 = -2.0000 mm, Y0 = 0.0000 mm 09:20:21 Integrated Voltage Measurement... 09:20:30 Voltage Offset Correction: v_off_ini = -102.729 uV, v_off_fin = -108.280 uV 09:20:30 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -27.907 uVs 09:20:30 Integrated Voltage Measurement... 09:20:39 Voltage Offset Correction: v_off_ini = -119.954 uV, v_off_fin = -101.771 uV 09:20:39 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 1.961 uVs 09:20:39 Integrated Voltage Measurement... 09:20:48 Voltage Offset Correction: v_off_ini = -104.712 uV, v_off_fin = -110.612 uV 09:20:48 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -22.759 uVs 09:20:48 Integrated Voltage Measurement... 09:20:57 Voltage Offset Correction: v_off_ini = -100.820 uV, v_off_fin = -97.462 uV 09:20:57 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 40.767 uVs 09:20:57 Integrated Voltage Measurement... 09:21:06 Voltage Offset Correction: v_off_ini = -100.787 uV, v_off_fin = -116.458 uV 09:21:06 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -12.314 uVs 09:21:06 Integrated Voltage Measurement... 09:21:15 Voltage Offset Correction: v_off_ini = -100.504 uV, v_off_fin = -96.612 uV 09:21:15 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 46.972 uVs 09:21:19 I1X Average and RMS of Measurements VT = -25.447 +- 7.484 uVs I1X = 5.089 +- 1.497 uTm 09:21:22 Wire Position: X0 = -2.0000 mm, Y0 = 0.0000 mm 09:21:26 Integrated Voltage Measurement... 09:21:35 Voltage Offset Correction: v_off_ini = -103.990 uV, v_off_fin = -114.014 uV 09:21:35 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -209.378 uVs 09:21:35 Integrated Voltage Measurement... 09:21:44 Voltage Offset Correction: v_off_ini = -104.914 uV, v_off_fin = -85.020 uV 09:21:44 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 202.434 uVs 09:21:44 Integrated Voltage Measurement... 09:21:53 Voltage Offset Correction: v_off_ini = -111.442 uV, v_off_fin = -111.409 uV 09:21:53 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -195.516 uVs 09:21:53 Integrated Voltage Measurement... 09:22:02 Voltage Offset Correction: v_off_ini = -96.552 uV, v_off_fin = -85.195 uV 09:22:02 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 201.751 uVs 09:22:02 Integrated Voltage Measurement... 09:22:11 Voltage Offset Correction: v_off_ini = -127.753 uV, v_off_fin = -117.272 uV 09:22:11 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -199.241 uVs 09:22:11 Integrated Voltage Measurement... 09:22:20 Voltage Offset Correction: v_off_ini = -93.159 uV, v_off_fin = -93.076 uV 09:22:20 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 204.023 uVs 09:22:24 I1Y Average and RMS of Measurements VT = -202.057 +- 2.984 uVs I1Y = -40.411 +- 0.597 uTm 09:22:29 Wire Position: X0 = -2.0000 mm, Y0 = 0.0000 mm 09:22:32 Integrated Voltage Measurement... 09:22:41 Voltage Offset Correction: v_off_ini = -84.024 uV, v_off_fin = -85.071 uV 09:22:41 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -37.449 uVs 09:22:41 Integrated Voltage Measurement... 09:22:50 Voltage Offset Correction: v_off_ini = -107.153 uV, v_off_fin = -95.836 uV 09:22:50 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 43.794 uVs 09:22:50 Integrated Voltage Measurement... 09:22:59 Voltage Offset Correction: v_off_ini = -87.528 uV, v_off_fin = -93.725 uV 09:22:59 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -18.908 uVs 09:22:59 Integrated Voltage Measurement... 09:23:08 Voltage Offset Correction: v_off_ini = -100.554 uV, v_off_fin = -99.985 uV 09:23:08 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 25.122 uVs 09:23:08 Integrated Voltage Measurement... 09:23:17 Voltage Offset Correction: v_off_ini = -95.434 uV, v_off_fin = -88.918 uV 09:23:17 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -28.694 uVs 09:23:17 Integrated Voltage Measurement... 09:23:26 Voltage Offset Correction: v_off_ini = -104.401 uV, v_off_fin = -90.916 uV 09:23:26 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 38.057 uVs 09:23:29 I2X Average and RMS of Measurements VT = -32.004 +- 7.658 uVs I2X = 2.944 +- 0.705 uTm^2 09:23:34 Wire Position: X0 = -2.0000 mm, Y0 = 0.0000 mm 09:23:36 Integrated Voltage Measurement... 09:23:45 Voltage Offset Correction: v_off_ini = -101.667 uV, v_off_fin = -100.687 uV 09:23:45 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -218.832 uVs 09:23:45 Integrated Voltage Measurement... 09:23:54 Voltage Offset Correction: v_off_ini = -87.120 uV, v_off_fin = -91.006 uV 09:23:54 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 194.144 uVs 09:23:55 Integrated Voltage Measurement... 09:24:03 Voltage Offset Correction: v_off_ini = -96.742 uV, v_off_fin = -110.242 uV 09:24:04 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -207.891 uVs 09:24:04 Integrated Voltage Measurement... 09:24:13 Voltage Offset Correction: v_off_ini = -103.357 uV, v_off_fin = -104.734 uV 09:24:13 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 211.697 uVs 09:24:13 Integrated Voltage Measurement... 09:24:22 Voltage Offset Correction: v_off_ini = -107.045 uV, v_off_fin = -112.622 uV 09:24:22 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -199.511 uVs 09:24:22 Integrated Voltage Measurement... 09:24:31 Voltage Offset Correction: v_off_ini = -105.574 uV, v_off_fin = -101.433 uV 09:24:31 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 212.475 uVs 09:24:33 I2Y Average and RMS of Measurements VT = -207.425 +- 1.687 uVs I2Y = -19.083 +- 0.155 uTm^2 09:24:38 Wire Position: X0 = -1.0000 mm, Y0 = 0.0000 mm 09:24:42 Integrated Voltage Measurement... 09:24:51 Voltage Offset Correction: v_off_ini = -108.796 uV, v_off_fin = -114.629 uV 09:24:51 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -39.873 uVs 09:24:51 Integrated Voltage Measurement... 09:25:00 Voltage Offset Correction: v_off_ini = -119.205 uV, v_off_fin = -103.680 uV 09:25:00 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 54.103 uVs 09:25:00 Integrated Voltage Measurement... 09:25:09 Voltage Offset Correction: v_off_ini = -100.769 uV, v_off_fin = -108.454 uV 09:25:09 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -40.134 uVs 09:25:09 Integrated Voltage Measurement... 09:25:18 Voltage Offset Correction: v_off_ini = -98.087 uV, v_off_fin = -105.627 uV 09:25:18 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 48.900 uVs 09:25:18 Integrated Voltage Measurement... 09:25:27 Voltage Offset Correction: v_off_ini = -111.067 uV, v_off_fin = -113.820 uV 09:25:27 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -53.641 uVs 09:25:27 Integrated Voltage Measurement... 09:25:36 Voltage Offset Correction: v_off_ini = -124.930 uV, v_off_fin = -107.696 uV 09:25:36 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 49.607 uVs 09:25:40 I1X Average and RMS of Measurements VT = -47.710 +- 2.946 uVs I1X = 9.542 +- 0.589 uTm 09:25:43 Wire Position: X0 = -1.0000 mm, Y0 = 0.0000 mm 09:25:47 Integrated Voltage Measurement... 09:25:56 Voltage Offset Correction: v_off_ini = -92.404 uV, v_off_fin = -107.462 uV 09:25:56 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -203.068 uVs 09:25:56 Integrated Voltage Measurement... 09:26:05 Voltage Offset Correction: v_off_ini = -88.796 uV, v_off_fin = -83.735 uV 09:26:05 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 202.571 uVs 09:26:05 Integrated Voltage Measurement... 09:26:14 Voltage Offset Correction: v_off_ini = -82.568 uV, v_off_fin = -103.583 uV 09:26:14 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -215.999 uVs 09:26:14 Integrated Voltage Measurement... 09:26:23 Voltage Offset Correction: v_off_ini = -72.592 uV, v_off_fin = -72.298 uV 09:26:23 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 198.565 uVs 09:26:23 Integrated Voltage Measurement... 09:26:32 Voltage Offset Correction: v_off_ini = -82.577 uV, v_off_fin = -100.940 uV 09:26:32 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -208.080 uVs 09:26:32 Integrated Voltage Measurement... 09:26:41 Voltage Offset Correction: v_off_ini = -82.447 uV, v_off_fin = -92.794 uV 09:26:41 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 185.906 uVs 09:26:45 I1Y Average and RMS of Measurements VT = -202.365 +- 4.213 uVs I1Y = -40.473 +- 0.843 uTm 09:26:50 Wire Position: X0 = -1.0000 mm, Y0 = 0.0000 mm 09:26:53 Integrated Voltage Measurement... 09:27:02 Voltage Offset Correction: v_off_ini = -75.709 uV, v_off_fin = -90.545 uV 09:27:02 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -66.146 uVs 09:27:02 Integrated Voltage Measurement... 09:27:11 Voltage Offset Correction: v_off_ini = -113.349 uV, v_off_fin = -97.301 uV 09:27:11 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 62.589 uVs 09:27:11 Integrated Voltage Measurement... 09:27:20 Voltage Offset Correction: v_off_ini = -93.222 uV, v_off_fin = -87.421 uV 09:27:20 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -45.206 uVs 09:27:20 Integrated Voltage Measurement... 09:27:29 Voltage Offset Correction: v_off_ini = -91.270 uV, v_off_fin = -104.248 uV 09:27:29 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 45.029 uVs 09:27:29 Integrated Voltage Measurement... 09:27:38 Voltage Offset Correction: v_off_ini = -84.524 uV, v_off_fin = -87.182 uV 09:27:38 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -38.617 uVs 09:27:38 Integrated Voltage Measurement... 09:27:47 Voltage Offset Correction: v_off_ini = -92.445 uV, v_off_fin = -98.433 uV 09:27:47 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 59.410 uVs 09:27:50 I2X Average and RMS of Measurements VT = -52.833 +- 8.310 uVs I2X = 4.861 +- 0.765 uTm^2 09:27:55 Wire Position: X0 = -1.0000 mm, Y0 = 0.0000 mm 09:27:57 Integrated Voltage Measurement... 09:28:06 Voltage Offset Correction: v_off_ini = -94.191 uV, v_off_fin = -96.960 uV 09:28:06 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -219.068 uVs 09:28:06 Integrated Voltage Measurement... 09:28:15 Voltage Offset Correction: v_off_ini = -95.223 uV, v_off_fin = -107.176 uV 09:28:15 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 235.525 uVs 09:28:15 Integrated Voltage Measurement... 09:28:24 Voltage Offset Correction: v_off_ini = -98.350 uV, v_off_fin = -100.301 uV 09:28:24 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = -221.972 uVs 09:28:24 Integrated Voltage Measurement... 09:28:33 Voltage Offset Correction: v_off_ini = -106.741 uV, v_off_fin = -96.507 uV 09:28:33 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 203.492 uVs 09:28:33 Integrated Voltage Measurement... 09:28:42 Voltage Offset Correction: v_off_ini = -99.932 uV, v_off_fin = -93.401 uV 09:28:43 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -218.902 uVs 09:28:43 Integrated Voltage Measurement... 09:28:52 Voltage Offset Correction: v_off_ini = -100.373 uV, v_off_fin = -95.879 uV 09:28:52 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 206.567 uVs 09:28:54 I2Y Average and RMS of Measurements VT = -217.588 +- 6.865 uVs I2Y = -20.018 +- 0.631 uTm^2 09:28:59 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:29:03 Integrated Voltage Measurement... 09:29:12 Voltage Offset Correction: v_off_ini = -102.801 uV, v_off_fin = -104.192 uV 09:29:12 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -81.818 uVs 09:29:12 Integrated Voltage Measurement... 09:29:21 Voltage Offset Correction: v_off_ini = -110.166 uV, v_off_fin = -100.585 uV 09:29:21 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 61.696 uVs 09:29:21 Integrated Voltage Measurement... 09:29:30 Voltage Offset Correction: v_off_ini = -110.488 uV, v_off_fin = -119.226 uV 09:29:30 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -73.786 uVs 09:29:30 Integrated Voltage Measurement... 09:29:39 Voltage Offset Correction: v_off_ini = -115.326 uV, v_off_fin = -104.796 uV 09:29:39 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 63.655 uVs 09:29:39 Integrated Voltage Measurement... 09:29:48 Voltage Offset Correction: v_off_ini = -120.096 uV, v_off_fin = -119.557 uV 09:29:48 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -70.007 uVs 09:29:48 Integrated Voltage Measurement... 09:29:57 Voltage Offset Correction: v_off_ini = -94.616 uV, v_off_fin = -98.132 uV 09:29:57 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 71.124 uVs 09:30:01 I1X Average and RMS of Measurements VT = -70.347 +- 1.249 uVs I1X = 14.069 +- 0.250 uTm 09:30:04 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:30:08 Integrated Voltage Measurement... 09:30:17 Voltage Offset Correction: v_off_ini = -132.103 uV, v_off_fin = -129.014 uV 09:30:17 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -220.707 uVs 09:30:17 Integrated Voltage Measurement... 09:30:26 Voltage Offset Correction: v_off_ini = -118.629 uV, v_off_fin = -103.091 uV 09:30:26 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 217.791 uVs 09:30:26 Integrated Voltage Measurement... 09:30:35 Voltage Offset Correction: v_off_ini = -118.092 uV, v_off_fin = -123.165 uV 09:30:35 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -219.438 uVs 09:30:35 Integrated Voltage Measurement... 09:30:44 Voltage Offset Correction: v_off_ini = -114.492 uV, v_off_fin = -108.144 uV 09:30:44 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 219.551 uVs 09:30:44 Integrated Voltage Measurement... 09:30:53 Voltage Offset Correction: v_off_ini = -122.828 uV, v_off_fin = -136.816 uV 09:30:53 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -212.356 uVs 09:30:53 Integrated Voltage Measurement... 09:31:02 Voltage Offset Correction: v_off_ini = -133.117 uV, v_off_fin = -112.939 uV 09:31:02 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 214.358 uVs 09:31:06 I1Y Average and RMS of Measurements VT = -217.367 +- 2.837 uVs I1Y = -43.473 +- 0.567 uTm 09:31:11 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:31:14 Integrated Voltage Measurement... 09:31:23 Voltage Offset Correction: v_off_ini = -111.494 uV, v_off_fin = -110.136 uV 09:31:23 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -65.319 uVs 09:31:23 Integrated Voltage Measurement... 09:31:32 Voltage Offset Correction: v_off_ini = -121.048 uV, v_off_fin = -140.718 uV 09:31:32 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 61.706 uVs 09:31:32 Integrated Voltage Measurement... 09:31:41 Voltage Offset Correction: v_off_ini = -118.875 uV, v_off_fin = -113.799 uV 09:31:41 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -79.237 uVs 09:31:41 Integrated Voltage Measurement... 09:31:50 Voltage Offset Correction: v_off_ini = -116.940 uV, v_off_fin = -131.707 uV 09:31:50 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 56.751 uVs 09:31:50 Integrated Voltage Measurement... 09:31:59 Voltage Offset Correction: v_off_ini = -120.430 uV, v_off_fin = -104.700 uV 09:31:59 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -88.183 uVs 09:31:59 Integrated Voltage Measurement... 09:32:08 Voltage Offset Correction: v_off_ini = -124.118 uV, v_off_fin = -125.870 uV 09:32:08 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 88.200 uVs 09:32:10 I2X Average and RMS of Measurements VT = -73.232 +- 10.734 uVs I2X = 6.737 +- 0.988 uTm^2 09:32:16 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:32:18 Integrated Voltage Measurement... 09:32:27 Voltage Offset Correction: v_off_ini = -115.795 uV, v_off_fin = -115.767 uV 09:32:27 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -205.440 uVs 09:32:27 Integrated Voltage Measurement... 09:32:36 Voltage Offset Correction: v_off_ini = -102.330 uV, v_off_fin = -109.368 uV 09:32:36 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 219.104 uVs 09:32:36 Integrated Voltage Measurement... 09:32:45 Voltage Offset Correction: v_off_ini = -108.649 uV, v_off_fin = -128.214 uV 09:32:45 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -192.825 uVs 09:32:45 Integrated Voltage Measurement... 09:32:54 Voltage Offset Correction: v_off_ini = -115.657 uV, v_off_fin = -101.407 uV 09:32:54 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 213.865 uVs 09:32:54 Integrated Voltage Measurement... 09:33:03 Voltage Offset Correction: v_off_ini = -110.424 uV, v_off_fin = -110.443 uV 09:33:03 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -228.334 uVs 09:33:03 Integrated Voltage Measurement... 09:33:12 Voltage Offset Correction: v_off_ini = -103.067 uV, v_off_fin = -100.501 uV 09:33:12 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 240.238 uVs 09:33:15 I2Y Average and RMS of Measurements VT = -216.634 +- 13.003 uVs I2Y = -19.930 +- 1.196 uTm^2 09:33:19 Wire Position: X0 = 1.0000 mm, Y0 = 0.0000 mm 09:33:24 Integrated Voltage Measurement... 09:33:33 Voltage Offset Correction: v_off_ini = -104.944 uV, v_off_fin = -105.763 uV 09:33:33 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -76.283 uVs 09:33:33 Integrated Voltage Measurement... 09:33:42 Voltage Offset Correction: v_off_ini = -90.934 uV, v_off_fin = -98.675 uV 09:33:42 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 98.267 uVs 09:33:42 Integrated Voltage Measurement... 09:33:51 Voltage Offset Correction: v_off_ini = -86.272 uV, v_off_fin = -99.410 uV 09:33:51 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -98.293 uVs 09:33:51 Integrated Voltage Measurement... 09:34:00 Voltage Offset Correction: v_off_ini = -115.154 uV, v_off_fin = -91.685 uV 09:34:00 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 79.876 uVs 09:34:00 Integrated Voltage Measurement... 09:34:09 Voltage Offset Correction: v_off_ini = -109.382 uV, v_off_fin = -109.510 uV 09:34:09 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -80.831 uVs 09:34:09 Integrated Voltage Measurement... 09:34:18 Voltage Offset Correction: v_off_ini = -93.405 uV, v_off_fin = -89.701 uV 09:34:18 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 65.322 uVs 09:34:22 I1X Average and RMS of Measurements VT = -83.145 +- 7.158 uVs I1X = 16.629 +- 1.432 uTm 09:34:25 Wire Position: X0 = 1.0000 mm, Y0 = 0.0000 mm 09:34:29 Integrated Voltage Measurement... 09:34:38 Voltage Offset Correction: v_off_ini = -104.173 uV, v_off_fin = -112.506 uV 09:34:38 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -212.396 uVs 09:34:38 Integrated Voltage Measurement... 09:34:47 Voltage Offset Correction: v_off_ini = -109.220 uV, v_off_fin = -90.018 uV 09:34:47 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 220.652 uVs 09:34:47 Integrated Voltage Measurement... 09:34:56 Voltage Offset Correction: v_off_ini = -99.912 uV, v_off_fin = -99.598 uV 09:34:56 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -212.383 uVs 09:34:56 Integrated Voltage Measurement... 09:35:05 Voltage Offset Correction: v_off_ini = -102.888 uV, v_off_fin = -98.399 uV 09:35:05 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 223.724 uVs 09:35:05 Integrated Voltage Measurement... 09:35:14 Voltage Offset Correction: v_off_ini = -95.179 uV, v_off_fin = -115.413 uV 09:35:14 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -231.262 uVs 09:35:14 Integrated Voltage Measurement... 09:35:23 Voltage Offset Correction: v_off_ini = -117.800 uV, v_off_fin = -112.856 uV 09:35:23 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 220.852 uVs 09:35:27 I1Y Average and RMS of Measurements VT = -220.212 +- 4.180 uVs I1Y = -44.042 +- 0.836 uTm 09:35:32 Wire Position: X0 = 1.0000 mm, Y0 = 0.0000 mm 09:35:35 Integrated Voltage Measurement... 09:35:44 Voltage Offset Correction: v_off_ini = -99.780 uV, v_off_fin = -106.400 uV 09:35:44 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -112.354 uVs 09:35:44 Integrated Voltage Measurement... 09:35:53 Voltage Offset Correction: v_off_ini = -107.745 uV, v_off_fin = -121.408 uV 09:35:53 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 96.394 uVs 09:35:53 Integrated Voltage Measurement... 09:36:02 Voltage Offset Correction: v_off_ini = -101.654 uV, v_off_fin = -99.051 uV 09:36:02 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -92.376 uVs 09:36:02 Integrated Voltage Measurement... 09:36:11 Voltage Offset Correction: v_off_ini = -108.123 uV, v_off_fin = -108.183 uV 09:36:11 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 81.862 uVs 09:36:11 Integrated Voltage Measurement... 09:36:20 Voltage Offset Correction: v_off_ini = -104.754 uV, v_off_fin = -114.337 uV 09:36:20 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -96.298 uVs 09:36:20 Integrated Voltage Measurement... 09:36:29 Voltage Offset Correction: v_off_ini = -118.409 uV, v_off_fin = -114.793 uV 09:36:29 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 79.361 uVs 09:36:31 I2X Average and RMS of Measurements VT = -93.108 +- 7.972 uVs I2X = 8.566 +- 0.733 uTm^2 09:36:36 Wire Position: X0 = 1.0000 mm, Y0 = 0.0000 mm 09:36:39 Integrated Voltage Measurement... 09:36:48 Voltage Offset Correction: v_off_ini = -91.499 uV, v_off_fin = -101.623 uV 09:36:48 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -223.963 uVs 09:36:48 Integrated Voltage Measurement... 09:36:57 Voltage Offset Correction: v_off_ini = -88.912 uV, v_off_fin = -97.126 uV 09:36:57 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 205.246 uVs 09:36:57 Integrated Voltage Measurement... 09:37:06 Voltage Offset Correction: v_off_ini = -118.954 uV, v_off_fin = -100.997 uV 09:37:06 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -223.420 uVs 09:37:06 Integrated Voltage Measurement... 09:37:15 Voltage Offset Correction: v_off_ini = -106.834 uV, v_off_fin = -112.816 uV 09:37:15 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 221.784 uVs 09:37:15 Integrated Voltage Measurement... 09:37:24 Voltage Offset Correction: v_off_ini = -105.767 uV, v_off_fin = -106.186 uV 09:37:24 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -238.184 uVs 09:37:24 Integrated Voltage Measurement... 09:37:33 Voltage Offset Correction: v_off_ini = -108.646 uV, v_off_fin = -86.835 uV 09:37:33 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 243.367 uVs 09:37:36 I2Y Average and RMS of Measurements VT = -225.994 +- 10.950 uVs I2Y = -20.791 +- 1.007 uTm^2 09:37:40 Wire Position: X0 = 2.0000 mm, Y0 = 0.0000 mm 09:37:45 Integrated Voltage Measurement... 09:37:54 Voltage Offset Correction: v_off_ini = -93.211 uV, v_off_fin = -94.360 uV 09:37:54 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -118.317 uVs 09:37:54 Integrated Voltage Measurement... 09:38:03 Voltage Offset Correction: v_off_ini = -85.642 uV, v_off_fin = -84.851 uV 09:38:03 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 107.055 uVs 09:38:03 Integrated Voltage Measurement... 09:38:12 Voltage Offset Correction: v_off_ini = -78.885 uV, v_off_fin = -94.312 uV 09:38:12 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -103.542 uVs 09:38:12 Integrated Voltage Measurement... 09:38:21 Voltage Offset Correction: v_off_ini = -90.374 uV, v_off_fin = -83.624 uV 09:38:21 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 101.014 uVs 09:38:21 Integrated Voltage Measurement... 09:38:30 Voltage Offset Correction: v_off_ini = -97.803 uV, v_off_fin = -99.184 uV 09:38:30 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -80.582 uVs 09:38:30 Integrated Voltage Measurement... 09:38:39 Voltage Offset Correction: v_off_ini = -84.263 uV, v_off_fin = -85.087 uV 09:38:39 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 107.979 uVs 09:38:43 I1X Average and RMS of Measurements VT = -103.082 +- 7.535 uVs I1X = 20.616 +- 1.507 uTm 09:38:45 Wire Position: X0 = 2.0000 mm, Y0 = 0.0000 mm 09:38:50 Integrated Voltage Measurement... 09:38:59 Voltage Offset Correction: v_off_ini = -88.187 uV, v_off_fin = -97.364 uV 09:38:59 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -229.422 uVs 09:38:59 Integrated Voltage Measurement... 09:39:08 Voltage Offset Correction: v_off_ini = -92.394 uV, v_off_fin = -84.821 uV 09:39:08 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 222.466 uVs 09:39:08 Integrated Voltage Measurement... 09:39:17 Voltage Offset Correction: v_off_ini = -94.552 uV, v_off_fin = -101.112 uV 09:39:17 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -227.313 uVs 09:39:17 Integrated Voltage Measurement... 09:39:26 Voltage Offset Correction: v_off_ini = -81.326 uV, v_off_fin = -68.709 uV 09:39:26 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 225.006 uVs 09:39:26 Integrated Voltage Measurement... 09:39:35 Voltage Offset Correction: v_off_ini = -82.554 uV, v_off_fin = -100.878 uV 09:39:35 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -217.741 uVs 09:39:35 Integrated Voltage Measurement... 09:39:44 Voltage Offset Correction: v_off_ini = -96.893 uV, v_off_fin = -90.992 uV 09:39:44 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 221.188 uVs 09:39:48 I1Y Average and RMS of Measurements VT = -223.856 +- 3.106 uVs I1Y = -44.771 +- 0.621 uTm 09:39:53 Wire Position: X0 = 2.0000 mm, Y0 = 0.0000 mm 09:39:56 Integrated Voltage Measurement... 09:40:05 Voltage Offset Correction: v_off_ini = -92.398 uV, v_off_fin = -77.752 uV 09:40:05 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -117.302 uVs 09:40:05 Integrated Voltage Measurement... 09:40:14 Voltage Offset Correction: v_off_ini = -95.815 uV, v_off_fin = -81.386 uV 09:40:14 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 114.438 uVs 09:40:14 Integrated Voltage Measurement... 09:40:23 Voltage Offset Correction: v_off_ini = -110.951 uV, v_off_fin = -84.003 uV 09:40:23 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -95.822 uVs 09:40:23 Integrated Voltage Measurement... 09:40:32 Voltage Offset Correction: v_off_ini = -95.495 uV, v_off_fin = -92.465 uV 09:40:32 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 98.971 uVs 09:40:32 Integrated Voltage Measurement... 09:40:41 Voltage Offset Correction: v_off_ini = -88.742 uV, v_off_fin = -88.556 uV 09:40:41 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -112.967 uVs 09:40:41 Integrated Voltage Measurement... 09:40:50 Voltage Offset Correction: v_off_ini = -89.504 uV, v_off_fin = -103.486 uV 09:40:50 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 137.315 uVs 09:40:52 I2X Average and RMS of Measurements VT = -112.802 +- 11.532 uVs I2X = 10.378 +- 1.061 uTm^2 09:40:57 Wire Position: X0 = 2.0000 mm, Y0 = 0.0000 mm 09:41:00 Integrated Voltage Measurement... 09:41:09 Voltage Offset Correction: v_off_ini = -85.019 uV, v_off_fin = -81.752 uV 09:41:09 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -221.718 uVs 09:41:09 Integrated Voltage Measurement... 09:41:18 Voltage Offset Correction: v_off_ini = -96.698 uV, v_off_fin = -97.847 uV 09:41:18 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 222.743 uVs 09:41:18 Integrated Voltage Measurement... 09:41:27 Voltage Offset Correction: v_off_ini = -81.502 uV, v_off_fin = -88.540 uV 09:41:27 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -221.983 uVs 09:41:27 Integrated Voltage Measurement... 09:41:36 Voltage Offset Correction: v_off_ini = -89.535 uV, v_off_fin = -62.812 uV 09:41:36 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 214.589 uVs 09:41:36 Integrated Voltage Measurement... 09:41:45 Voltage Offset Correction: v_off_ini = -91.198 uV, v_off_fin = -78.895 uV 09:41:45 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -205.712 uVs 09:41:45 Integrated Voltage Measurement... 09:41:54 Voltage Offset Correction: v_off_ini = -60.784 uV, v_off_fin = -77.686 uV 09:41:54 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 197.089 uVs 09:41:57 I2Y Average and RMS of Measurements VT = -213.972 +- 9.034 uVs I2Y = -19.685 +- 0.831 uTm^2 09:42:04 Wire Position: X0 = 0.0000 mm, Y0 = -1.0000 mm 09:42:09 Integrated Voltage Measurement... 09:42:18 Voltage Offset Correction: v_off_ini = -98.340 uV, v_off_fin = -87.523 uV 09:42:18 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -61.240 uVs 09:42:18 Integrated Voltage Measurement... 09:42:27 Voltage Offset Correction: v_off_ini = -86.023 uV, v_off_fin = -87.531 uV 09:42:27 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 71.298 uVs 09:42:27 Integrated Voltage Measurement... 09:42:36 Voltage Offset Correction: v_off_ini = -89.230 uV, v_off_fin = -93.387 uV 09:42:36 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -79.453 uVs 09:42:36 Integrated Voltage Measurement... 09:42:45 Voltage Offset Correction: v_off_ini = -92.020 uV, v_off_fin = -86.387 uV 09:42:45 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 85.739 uVs 09:42:45 Integrated Voltage Measurement... 09:42:54 Voltage Offset Correction: v_off_ini = -77.551 uV, v_off_fin = -84.334 uV 09:42:54 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -81.758 uVs 09:42:54 Integrated Voltage Measurement... 09:43:03 Voltage Offset Correction: v_off_ini = -70.564 uV, v_off_fin = -71.103 uV 09:43:03 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 75.073 uVs 09:43:07 I1X Average and RMS of Measurements VT = -75.760 +- 6.925 uVs I1X = 15.152 +- 1.385 uTm 09:43:09 Wire Position: X0 = 0.0000 mm, Y0 = -1.0000 mm 09:43:14 Integrated Voltage Measurement... 09:43:23 Voltage Offset Correction: v_off_ini = -70.799 uV, v_off_fin = -89.388 uV 09:43:23 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -217.385 uVs 09:43:23 Integrated Voltage Measurement... 09:43:32 Voltage Offset Correction: v_off_ini = -74.901 uV, v_off_fin = -65.072 uV 09:43:32 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 199.229 uVs 09:43:32 Integrated Voltage Measurement... 09:43:41 Voltage Offset Correction: v_off_ini = -86.500 uV, v_off_fin = -78.674 uV 09:43:41 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -205.292 uVs 09:43:41 Integrated Voltage Measurement... 09:43:50 Voltage Offset Correction: v_off_ini = -66.505 uV, v_off_fin = -54.725 uV 09:43:50 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 197.604 uVs 09:43:50 Integrated Voltage Measurement... 09:43:59 Voltage Offset Correction: v_off_ini = -84.181 uV, v_off_fin = -85.621 uV 09:43:59 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -213.513 uVs 09:43:59 Integrated Voltage Measurement... 09:44:08 Voltage Offset Correction: v_off_ini = -60.335 uV, v_off_fin = -60.918 uV 09:44:08 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 204.320 uVs 09:44:12 I1Y Average and RMS of Measurements VT = -206.224 +- 3.386 uVs I1Y = -41.245 +- 0.677 uTm 09:44:14 Wire Position: X0 = 0.0000 mm, Y0 = -1.0000 mm 09:44:17 Integrated Voltage Measurement... 09:44:26 Voltage Offset Correction: v_off_ini = -73.434 uV, v_off_fin = -70.472 uV 09:44:26 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -81.789 uVs 09:44:26 Integrated Voltage Measurement... 09:44:35 Voltage Offset Correction: v_off_ini = -81.266 uV, v_off_fin = -89.995 uV 09:44:35 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 44.801 uVs 09:44:35 Integrated Voltage Measurement... 09:44:44 Voltage Offset Correction: v_off_ini = -85.940 uV, v_off_fin = -68.192 uV 09:44:44 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -52.668 uVs 09:44:44 Integrated Voltage Measurement... 09:44:53 Voltage Offset Correction: v_off_ini = -65.923 uV, v_off_fin = -68.093 uV 09:44:53 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 61.902 uVs 09:44:53 Integrated Voltage Measurement... 09:45:02 Voltage Offset Correction: v_off_ini = -63.888 uV, v_off_fin = -70.419 uV 09:45:02 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -70.315 uVs 09:45:02 Integrated Voltage Measurement... 09:45:11 Voltage Offset Correction: v_off_ini = -73.338 uV, v_off_fin = -62.799 uV 09:45:11 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 70.252 uVs 09:45:14 I2X Average and RMS of Measurements VT = -63.621 +- 5.312 uVs I2X = 5.853 +- 0.489 uTm^2 09:45:16 Wire Position: X0 = 0.0000 mm, Y0 = -1.0000 mm 09:45:19 Integrated Voltage Measurement... 09:45:28 Voltage Offset Correction: v_off_ini = -53.537 uV, v_off_fin = -58.942 uV 09:45:28 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -207.683 uVs 09:45:28 Integrated Voltage Measurement... 09:45:37 Voltage Offset Correction: v_off_ini = -64.485 uV, v_off_fin = -53.932 uV 09:45:37 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 219.449 uVs 09:45:37 Integrated Voltage Measurement... 09:45:46 Voltage Offset Correction: v_off_ini = -66.088 uV, v_off_fin = -89.947 uV 09:45:46 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -222.088 uVs 09:45:46 Integrated Voltage Measurement... 09:45:55 Voltage Offset Correction: v_off_ini = -82.684 uV, v_off_fin = -66.183 uV 09:45:55 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 212.072 uVs 09:45:55 Integrated Voltage Measurement... 09:46:04 Voltage Offset Correction: v_off_ini = -69.046 uV, v_off_fin = -64.834 uV 09:46:04 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -202.327 uVs 09:46:04 Integrated Voltage Measurement... 09:46:13 Voltage Offset Correction: v_off_ini = -50.907 uV, v_off_fin = -61.183 uV 09:46:13 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 213.922 uVs 09:46:15 I2Y Average and RMS of Measurements VT = -212.923 +- 3.684 uVs I2Y = -19.589 +- 0.339 uTm^2 09:46:19 Wire Position: X0 = 0.0000 mm, Y0 = -0.5000 mm 09:46:24 Integrated Voltage Measurement... 09:46:33 Voltage Offset Correction: v_off_ini = -61.917 uV, v_off_fin = -59.143 uV 09:46:33 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.4999 mm dXY = -0.4999 mm VT = -64.876 uVs 09:46:33 Integrated Voltage Measurement... 09:46:42 Voltage Offset Correction: v_off_ini = -59.358 uV, v_off_fin = -47.827 uV 09:46:42 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 77.150 uVs 09:46:42 Integrated Voltage Measurement... 09:46:51 Voltage Offset Correction: v_off_ini = -52.663 uV, v_off_fin = -54.359 uV 09:46:51 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -69.797 uVs 09:46:51 Integrated Voltage Measurement... 09:47:00 Voltage Offset Correction: v_off_ini = -51.614 uV, v_off_fin = -56.589 uV 09:47:00 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 75.188 uVs 09:47:00 Integrated Voltage Measurement... 09:47:09 Voltage Offset Correction: v_off_ini = -64.227 uV, v_off_fin = -67.298 uV 09:47:09 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -86.332 uVs 09:47:09 Integrated Voltage Measurement... 09:47:18 Voltage Offset Correction: v_off_ini = -46.075 uV, v_off_fin = -40.668 uV 09:47:18 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 56.888 uVs 09:47:22 I1X Average and RMS of Measurements VT = -71.705 +- 0.608 uVs I1X = 14.341 +- 0.121 uTm 09:47:24 Wire Position: X0 = 0.0000 mm, Y0 = -0.5000 mm 09:47:29 Integrated Voltage Measurement... 09:47:38 Voltage Offset Correction: v_off_ini = -55.495 uV, v_off_fin = -75.449 uV 09:47:38 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -197.637 uVs 09:47:38 Integrated Voltage Measurement... 09:47:47 Voltage Offset Correction: v_off_ini = -48.282 uV, v_off_fin = -37.620 uV 09:47:47 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 206.029 uVs 09:47:47 Integrated Voltage Measurement... 09:47:56 Voltage Offset Correction: v_off_ini = -42.113 uV, v_off_fin = -63.042 uV 09:47:56 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -222.061 uVs 09:47:56 Integrated Voltage Measurement... 09:48:05 Voltage Offset Correction: v_off_ini = -46.905 uV, v_off_fin = -40.201 uV 09:48:05 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 210.630 uVs 09:48:05 Integrated Voltage Measurement... 09:48:14 Voltage Offset Correction: v_off_ini = -51.235 uV, v_off_fin = -47.490 uV 09:48:14 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -227.895 uVs 09:48:14 Integrated Voltage Measurement... 09:48:23 Voltage Offset Correction: v_off_ini = -37.511 uV, v_off_fin = -19.126 uV 09:48:23 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 211.526 uVs 09:48:27 I1Y Average and RMS of Measurements VT = -212.630 +- 7.757 uVs I1Y = -42.527 +- 1.550 uTm 09:48:29 Wire Position: X0 = 0.0000 mm, Y0 = -0.5000 mm 09:48:32 Integrated Voltage Measurement... 09:48:41 Voltage Offset Correction: v_off_ini = -47.177 uV, v_off_fin = -41.797 uV 09:48:41 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -84.316 uVs 09:48:41 Integrated Voltage Measurement... 09:48:50 Voltage Offset Correction: v_off_ini = -62.661 uV, v_off_fin = -53.798 uV 09:48:50 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 86.229 uVs 09:48:50 Integrated Voltage Measurement... 09:48:59 Voltage Offset Correction: v_off_ini = -43.328 uV, v_off_fin = -47.797 uV 09:48:59 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -85.990 uVs 09:48:59 Integrated Voltage Measurement... 09:49:08 Voltage Offset Correction: v_off_ini = -34.735 uV, v_off_fin = -52.814 uV 09:49:08 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 61.068 uVs 09:49:08 Integrated Voltage Measurement... 09:49:17 Voltage Offset Correction: v_off_ini = -63.070 uV, v_off_fin = -43.893 uV 09:49:17 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -57.902 uVs 09:49:17 Integrated Voltage Measurement... 09:49:26 Voltage Offset Correction: v_off_ini = -41.489 uV, v_off_fin = -45.209 uV 09:49:26 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 79.466 uVs 09:49:29 I2X Average and RMS of Measurements VT = -75.828 +- 6.965 uVs I2X = 6.976 +- 0.641 uTm^2 09:49:31 Wire Position: X0 = 0.0000 mm, Y0 = -0.5000 mm 09:49:34 Integrated Voltage Measurement... 09:49:43 Voltage Offset Correction: v_off_ini = -41.088 uV, v_off_fin = -43.650 uV 09:49:43 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -211.950 uVs 09:49:43 Integrated Voltage Measurement... 09:49:52 Voltage Offset Correction: v_off_ini = -37.491 uV, v_off_fin = -40.983 uV 09:49:52 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 220.204 uVs 09:49:52 Integrated Voltage Measurement... 09:50:01 Voltage Offset Correction: v_off_ini = -34.278 uV, v_off_fin = -34.442 uV 09:50:01 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -220.740 uVs 09:50:01 Integrated Voltage Measurement... 09:50:10 Voltage Offset Correction: v_off_ini = -46.302 uV, v_off_fin = -45.859 uV 09:50:10 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 242.735 uVs 09:50:10 Integrated Voltage Measurement... 09:50:19 Voltage Offset Correction: v_off_ini = -37.837 uV, v_off_fin = -37.480 uV 09:50:19 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -226.144 uVs 09:50:19 Integrated Voltage Measurement... 09:50:28 Voltage Offset Correction: v_off_ini = -38.312 uV, v_off_fin = -46.846 uV 09:50:28 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 219.162 uVs 09:50:30 I2Y Average and RMS of Measurements VT = -223.489 +- 6.421 uVs I2Y = -20.561 +- 0.591 uTm^2 09:50:34 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:50:39 Integrated Voltage Measurement... 09:50:48 Voltage Offset Correction: v_off_ini = -51.525 uV, v_off_fin = -45.118 uV 09:50:48 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -58.134 uVs 09:50:48 Integrated Voltage Measurement... 09:50:57 Voltage Offset Correction: v_off_ini = -50.347 uV, v_off_fin = -56.786 uV 09:50:57 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 64.350 uVs 09:50:57 Integrated Voltage Measurement... 09:51:06 Voltage Offset Correction: v_off_ini = -63.484 uV, v_off_fin = -65.456 uV 09:51:06 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -59.751 uVs 09:51:06 Integrated Voltage Measurement... 09:51:15 Voltage Offset Correction: v_off_ini = -45.074 uV, v_off_fin = -45.595 uV 09:51:15 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 72.376 uVs 09:51:15 Integrated Voltage Measurement... 09:51:24 Voltage Offset Correction: v_off_ini = -70.393 uV, v_off_fin = -66.770 uV 09:51:24 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -60.997 uVs 09:51:24 Integrated Voltage Measurement... 09:51:33 Voltage Offset Correction: v_off_ini = -65.849 uV, v_off_fin = -62.155 uV 09:51:33 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 65.254 uVs 09:51:37 I1X Average and RMS of Measurements VT = -63.477 +- 1.984 uVs I1X = 12.695 +- 0.397 uTm 09:51:40 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:51:44 Integrated Voltage Measurement... 09:51:53 Voltage Offset Correction: v_off_ini = -45.359 uV, v_off_fin = -64.768 uV 09:51:53 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -228.270 uVs 09:51:53 Integrated Voltage Measurement... 09:52:02 Voltage Offset Correction: v_off_ini = -48.989 uV, v_off_fin = -46.067 uV 09:52:02 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 241.394 uVs 09:52:02 Integrated Voltage Measurement... 09:52:11 Voltage Offset Correction: v_off_ini = -58.129 uV, v_off_fin = -67.605 uV 09:52:11 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -226.818 uVs 09:52:11 Integrated Voltage Measurement... 09:52:20 Voltage Offset Correction: v_off_ini = -69.909 uV, v_off_fin = -64.431 uV 09:52:20 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 215.966 uVs 09:52:20 Integrated Voltage Measurement... 09:52:29 Voltage Offset Correction: v_off_ini = -69.669 uV, v_off_fin = -63.720 uV 09:52:29 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -221.543 uVs 09:52:29 Integrated Voltage Measurement... 09:52:38 Voltage Offset Correction: v_off_ini = -66.773 uV, v_off_fin = -40.737 uV 09:52:38 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 215.740 uVs 09:52:42 I1Y Average and RMS of Measurements VT = -224.955 +- 7.074 uVs I1Y = -44.990 +- 1.414 uTm 09:52:44 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:52:47 Integrated Voltage Measurement... 09:52:56 Voltage Offset Correction: v_off_ini = -57.814 uV, v_off_fin = -83.719 uV 09:52:56 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -58.801 uVs 09:52:56 Integrated Voltage Measurement... 09:53:05 Voltage Offset Correction: v_off_ini = -66.660 uV, v_off_fin = -73.229 uV 09:53:05 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 83.644 uVs 09:53:05 Integrated Voltage Measurement... 09:53:14 Voltage Offset Correction: v_off_ini = -59.867 uV, v_off_fin = -54.843 uV 09:53:14 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -74.910 uVs 09:53:14 Integrated Voltage Measurement... 09:53:23 Voltage Offset Correction: v_off_ini = -58.332 uV, v_off_fin = -72.823 uV 09:53:23 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 67.007 uVs 09:53:23 Integrated Voltage Measurement... 09:53:32 Voltage Offset Correction: v_off_ini = -68.808 uV, v_off_fin = -55.122 uV 09:53:32 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -67.610 uVs 09:53:32 Integrated Voltage Measurement... 09:53:41 Voltage Offset Correction: v_off_ini = -60.916 uV, v_off_fin = -63.299 uV 09:53:41 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 79.539 uVs 09:53:44 I2X Average and RMS of Measurements VT = -71.919 +- 1.176 uVs I2X = 6.617 +- 0.108 uTm^2 09:53:46 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:53:49 Integrated Voltage Measurement... 09:53:58 Voltage Offset Correction: v_off_ini = -62.044 uV, v_off_fin = -64.273 uV 09:53:58 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -206.534 uVs 09:53:58 Integrated Voltage Measurement... 09:54:07 Voltage Offset Correction: v_off_ini = -63.158 uV, v_off_fin = -73.744 uV 09:54:07 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 227.078 uVs 09:54:07 Integrated Voltage Measurement... 09:54:16 Voltage Offset Correction: v_off_ini = -61.759 uV, v_off_fin = -59.869 uV 09:54:16 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -235.895 uVs 09:54:16 Integrated Voltage Measurement... 09:54:25 Voltage Offset Correction: v_off_ini = -57.969 uV, v_off_fin = -51.211 uV 09:54:25 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 226.398 uVs 09:54:25 Integrated Voltage Measurement... 09:54:34 Voltage Offset Correction: v_off_ini = -44.617 uV, v_off_fin = -53.863 uV 09:54:34 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -223.000 uVs 09:54:34 Integrated Voltage Measurement... 09:54:43 Voltage Offset Correction: v_off_ini = -43.993 uV, v_off_fin = -35.309 uV 09:54:43 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 201.333 uVs 09:54:45 I2Y Average and RMS of Measurements VT = -220.040 +- 8.079 uVs I2Y = -20.244 +- 0.743 uTm^2 09:54:49 Wire Position: X0 = 0.0000 mm, Y0 = 0.5000 mm 09:54:54 Integrated Voltage Measurement... 09:55:03 Voltage Offset Correction: v_off_ini = -68.763 uV, v_off_fin = -63.502 uV 09:55:03 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -72.349 uVs 09:55:03 Integrated Voltage Measurement... 09:55:12 Voltage Offset Correction: v_off_ini = -92.519 uV, v_off_fin = -72.159 uV 09:55:12 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 46.778 uVs 09:55:12 Integrated Voltage Measurement... 09:55:21 Voltage Offset Correction: v_off_ini = -65.240 uV, v_off_fin = -72.360 uV 09:55:21 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -57.015 uVs 09:55:21 Integrated Voltage Measurement... 09:55:30 Voltage Offset Correction: v_off_ini = -69.577 uV, v_off_fin = -52.832 uV 09:55:30 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = 55.523 uVs 09:55:30 Integrated Voltage Measurement... 09:55:39 Voltage Offset Correction: v_off_ini = -55.466 uV, v_off_fin = -63.431 uV 09:55:39 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -68.619 uVs 09:55:39 Integrated Voltage Measurement... 09:55:48 Voltage Offset Correction: v_off_ini = -54.539 uV, v_off_fin = -60.344 uV 09:55:48 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 43.828 uVs 09:55:52 I1X Average and RMS of Measurements VT = -57.352 +- 1.564 uVs I1X = 11.470 +- 0.313 uTm 09:55:55 Wire Position: X0 = 0.0000 mm, Y0 = 0.5000 mm 09:55:59 Integrated Voltage Measurement... 09:56:08 Voltage Offset Correction: v_off_ini = -72.028 uV, v_off_fin = -90.472 uV 09:56:08 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = -225.335 uVs 09:56:08 Integrated Voltage Measurement... 09:56:17 Voltage Offset Correction: v_off_ini = -69.224 uV, v_off_fin = -53.914 uV 09:56:17 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 230.351 uVs 09:56:17 Integrated Voltage Measurement... 09:56:26 Voltage Offset Correction: v_off_ini = -74.478 uV, v_off_fin = -89.283 uV 09:56:26 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -203.204 uVs 09:56:26 Integrated Voltage Measurement... 09:56:35 Voltage Offset Correction: v_off_ini = -88.227 uV, v_off_fin = -66.002 uV 09:56:35 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 221.728 uVs 09:56:35 Integrated Voltage Measurement... 09:56:44 Voltage Offset Correction: v_off_ini = -94.820 uV, v_off_fin = -92.345 uV 09:56:44 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -233.373 uVs 09:56:44 Integrated Voltage Measurement... 09:56:53 Voltage Offset Correction: v_off_ini = -71.172 uV, v_off_fin = -72.191 uV 09:56:53 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 212.843 uVs 09:56:57 I1Y Average and RMS of Measurements VT = -221.139 +- 6.430 uVs I1Y = -44.227 +- 1.287 uTm 09:57:00 Wire Position: X0 = 0.0000 mm, Y0 = 0.5000 mm 09:57:02 Integrated Voltage Measurement... 09:57:11 Voltage Offset Correction: v_off_ini = -78.062 uV, v_off_fin = -88.821 uV 09:57:11 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -77.457 uVs 09:57:11 Integrated Voltage Measurement... 09:57:20 Voltage Offset Correction: v_off_ini = -94.185 uV, v_off_fin = -95.931 uV 09:57:20 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 90.807 uVs 09:57:20 Integrated Voltage Measurement... 09:57:29 Voltage Offset Correction: v_off_ini = -65.237 uV, v_off_fin = -86.495 uV 09:57:29 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -53.713 uVs 09:57:29 Integrated Voltage Measurement... 09:57:38 Voltage Offset Correction: v_off_ini = -94.717 uV, v_off_fin = -97.208 uV 09:57:38 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 84.736 uVs 09:57:38 Integrated Voltage Measurement... 09:57:47 Voltage Offset Correction: v_off_ini = -93.840 uV, v_off_fin = -93.315 uV 09:57:47 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -66.952 uVs 09:57:47 Integrated Voltage Measurement... 09:57:56 Voltage Offset Correction: v_off_ini = -105.055 uV, v_off_fin = -108.878 uV 09:57:56 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 70.445 uVs 09:57:59 I2X Average and RMS of Measurements VT = -74.018 +- 7.155 uVs I2X = 6.810 +- 0.658 uTm^2 09:58:01 Wire Position: X0 = 0.0000 mm, Y0 = 0.5000 mm 09:58:04 Integrated Voltage Measurement... 09:58:13 Voltage Offset Correction: v_off_ini = -95.772 uV, v_off_fin = -109.379 uV 09:58:13 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -219.762 uVs 09:58:13 Integrated Voltage Measurement... 09:58:22 Voltage Offset Correction: v_off_ini = -102.984 uV, v_off_fin = -98.483 uV 09:58:22 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 230.773 uVs 09:58:22 Integrated Voltage Measurement... 09:58:31 Voltage Offset Correction: v_off_ini = -99.283 uV, v_off_fin = -108.797 uV 09:58:31 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -245.835 uVs 09:58:31 Integrated Voltage Measurement... 09:58:40 Voltage Offset Correction: v_off_ini = -105.714 uV, v_off_fin = -89.758 uV 09:58:40 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 225.735 uVs 09:58:40 Integrated Voltage Measurement... 09:58:49 Voltage Offset Correction: v_off_ini = -102.262 uV, v_off_fin = -96.183 uV 09:58:49 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -228.963 uVs 09:58:49 Integrated Voltage Measurement... 09:58:58 Voltage Offset Correction: v_off_ini = -99.911 uV, v_off_fin = -97.576 uV 09:58:58 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 237.846 uVs 09:59:00 I2Y Average and RMS of Measurements VT = -231.486 +- 4.503 uVs I2Y = -21.297 +- 0.414 uTm^2 09:59:05 Wire Position: X0 = 0.0000 mm, Y0 = 1.0000 mm 09:59:09 Integrated Voltage Measurement... 09:59:18 Voltage Offset Correction: v_off_ini = -112.783 uV, v_off_fin = -104.488 uV 09:59:18 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -64.746 uVs 09:59:18 Integrated Voltage Measurement... 09:59:27 Voltage Offset Correction: v_off_ini = -96.406 uV, v_off_fin = -100.211 uV 09:59:27 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 44.861 uVs 09:59:27 Integrated Voltage Measurement... 09:59:36 Voltage Offset Correction: v_off_ini = -93.197 uV, v_off_fin = -105.291 uV 09:59:36 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -45.248 uVs 09:59:36 Integrated Voltage Measurement... 09:59:45 Voltage Offset Correction: v_off_ini = -100.353 uV, v_off_fin = -98.551 uV 09:59:45 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 87.076 uVs 09:59:45 Integrated Voltage Measurement... 09:59:54 Voltage Offset Correction: v_off_ini = -92.513 uV, v_off_fin = -103.608 uV 09:59:54 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -64.825 uVs 09:59:54 Integrated Voltage Measurement... 10:00:03 Voltage Offset Correction: v_off_ini = -98.968 uV, v_off_fin = -83.755 uV 10:00:03 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 33.124 uVs 10:00:07 I1X Average and RMS of Measurements VT = -56.647 +- 7.137 uVs I1X = 11.329 +- 1.427 uTm 10:00:10 Wire Position: X0 = 0.0000 mm, Y0 = 1.0000 mm 10:00:14 Integrated Voltage Measurement... 10:00:23 Voltage Offset Correction: v_off_ini = -78.385 uV, v_off_fin = -93.839 uV 10:00:23 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = -228.606 uVs 10:00:23 Integrated Voltage Measurement... 10:00:32 Voltage Offset Correction: v_off_ini = -81.776 uV, v_off_fin = -66.695 uV 10:00:32 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 240.774 uVs 10:00:32 Integrated Voltage Measurement... 10:00:41 Voltage Offset Correction: v_off_ini = -72.774 uV, v_off_fin = -89.985 uV 10:00:41 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -224.884 uVs 10:00:41 Integrated Voltage Measurement... 10:00:50 Voltage Offset Correction: v_off_ini = -95.530 uV, v_off_fin = -72.335 uV 10:00:50 Integrated Voltage Measurement dXYus = -0.5001 mm, dXYds = -0.5001 mm dXY = -0.5001 mm VT = 240.482 uVs 10:00:50 Integrated Voltage Measurement... 10:00:59 Voltage Offset Correction: v_off_ini = -84.357 uV, v_off_fin = -106.479 uV 10:00:59 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -227.877 uVs 10:00:59 Integrated Voltage Measurement... 10:01:08 Voltage Offset Correction: v_off_ini = -82.966 uV, v_off_fin = -68.794 uV 10:01:08 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 247.308 uVs 10:01:12 I1Y Average and RMS of Measurements VT = -234.988 +- 2.015 uVs I1Y = -46.997 +- 0.405 uTm 10:01:15 Wire Position: X0 = 0.0000 mm, Y0 = 1.0000 mm 10:01:17 Integrated Voltage Measurement... 10:01:26 Voltage Offset Correction: v_off_ini = -78.684 uV, v_off_fin = -75.970 uV 10:01:26 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -61.286 uVs 10:01:26 Integrated Voltage Measurement... 10:01:35 Voltage Offset Correction: v_off_ini = -69.936 uV, v_off_fin = -63.159 uV 10:01:35 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 57.907 uVs 10:01:35 Integrated Voltage Measurement... 10:01:44 Voltage Offset Correction: v_off_ini = -73.436 uV, v_off_fin = -67.959 uV 10:01:44 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -84.908 uVs 10:01:44 Integrated Voltage Measurement... 10:01:53 Voltage Offset Correction: v_off_ini = -73.284 uV, v_off_fin = -81.990 uV 10:01:53 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 56.920 uVs 10:01:53 Integrated Voltage Measurement... 10:02:02 Voltage Offset Correction: v_off_ini = -58.229 uV, v_off_fin = -64.274 uV 10:02:02 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -48.949 uVs 10:02:02 Integrated Voltage Measurement... 10:02:11 Voltage Offset Correction: v_off_ini = -68.964 uV, v_off_fin = -69.775 uV 10:02:11 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 52.007 uVs 10:02:14 I2X Average and RMS of Measurements VT = -60.329 +- 8.359 uVs I2X = 5.550 +- 0.769 uTm^2 10:02:16 Wire Position: X0 = 0.0000 mm, Y0 = 1.0000 mm 10:02:19 Integrated Voltage Measurement... 10:02:28 Voltage Offset Correction: v_off_ini = -57.326 uV, v_off_fin = -66.324 uV 10:02:28 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -228.838 uVs 10:02:28 Integrated Voltage Measurement... 10:02:37 Voltage Offset Correction: v_off_ini = -74.343 uV, v_off_fin = -67.112 uV 10:02:37 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 223.905 uVs 10:02:37 Integrated Voltage Measurement... 10:02:46 Voltage Offset Correction: v_off_ini = -52.009 uV, v_off_fin = -57.223 uV 10:02:46 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -248.585 uVs 10:02:46 Integrated Voltage Measurement... 10:02:55 Voltage Offset Correction: v_off_ini = -73.256 uV, v_off_fin = -64.001 uV 10:02:55 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 247.969 uVs 10:02:55 Integrated Voltage Measurement... 10:03:04 Voltage Offset Correction: v_off_ini = -80.771 uV, v_off_fin = -62.221 uV 10:03:04 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -241.428 uVs 10:03:04 Integrated Voltage Measurement... 10:03:13 Voltage Offset Correction: v_off_ini = -59.158 uV, v_off_fin = -56.012 uV 10:03:13 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 242.089 uVs 10:03:16 I2Y Average and RMS of Measurements VT = -238.802 +- 9.184 uVs I2Y = -21.970 +- 0.845 uTm^2 10:03:32 Wire Position: X0 = -2.0000 mm, Y0 = 0.0000 mm 10:03:36 Integrated Voltage Measurement... 10:03:45 Voltage Offset Correction: v_off_ini = -64.726 uV, v_off_fin = -62.827 uV 10:03:45 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -33.691 uVs 10:03:45 Integrated Voltage Measurement... 10:03:54 Voltage Offset Correction: v_off_ini = -65.244 uV, v_off_fin = -58.741 uV 10:03:54 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 35.217 uVs 10:03:54 Integrated Voltage Measurement... 10:04:03 Voltage Offset Correction: v_off_ini = -58.852 uV, v_off_fin = -55.111 uV 10:04:03 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -28.658 uVs 10:04:03 Integrated Voltage Measurement... 10:04:12 Voltage Offset Correction: v_off_ini = -47.776 uV, v_off_fin = -33.900 uV 10:04:12 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 19.858 uVs 10:04:12 Integrated Voltage Measurement... 10:04:21 Voltage Offset Correction: v_off_ini = -46.466 uV, v_off_fin = -65.618 uV 10:04:21 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -10.020 uVs 10:04:21 Integrated Voltage Measurement... 10:04:30 Voltage Offset Correction: v_off_ini = -60.141 uV, v_off_fin = -61.681 uV 10:04:30 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 36.986 uVs 10:04:34 I1X Average and RMS of Measurements VT = -27.405 +- 4.994 uVs I1X = 5.481 +- 0.999 uTm 10:04:37 Wire Position: X0 = -2.0000 mm, Y0 = 0.0000 mm 10:04:41 Integrated Voltage Measurement... 10:04:50 Voltage Offset Correction: v_off_ini = -64.973 uV, v_off_fin = -81.376 uV 10:04:50 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -189.557 uVs 10:04:50 Integrated Voltage Measurement... 10:04:59 Voltage Offset Correction: v_off_ini = -69.259 uV, v_off_fin = -62.478 uV 10:04:59 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 172.059 uVs 10:04:59 Integrated Voltage Measurement... 10:05:08 Voltage Offset Correction: v_off_ini = -73.295 uV, v_off_fin = -78.600 uV 10:05:08 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -186.163 uVs 10:05:08 Integrated Voltage Measurement... 10:05:17 Voltage Offset Correction: v_off_ini = -67.014 uV, v_off_fin = -57.229 uV 10:05:17 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 175.631 uVs 10:05:17 Integrated Voltage Measurement... 10:05:26 Voltage Offset Correction: v_off_ini = -67.268 uV, v_off_fin = -69.003 uV 10:05:26 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.4999 mm dXY = 0.4999 mm VT = -186.119 uVs 10:05:26 Integrated Voltage Measurement... 10:05:35 Voltage Offset Correction: v_off_ini = -67.898 uV, v_off_fin = -58.286 uV 10:05:35 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 189.495 uVs 10:05:39 I1Y Average and RMS of Measurements VT = -183.171 +- 3.278 uVs I1Y = -36.635 +- 0.657 uTm 10:05:44 Wire Position: X0 = -2.0000 mm, Y0 = 0.0000 mm 10:05:47 Integrated Voltage Measurement... 10:05:56 Voltage Offset Correction: v_off_ini = -88.238 uV, v_off_fin = -73.403 uV 10:05:56 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -28.537 uVs 10:05:56 Integrated Voltage Measurement... 10:06:05 Voltage Offset Correction: v_off_ini = -71.290 uV, v_off_fin = -59.406 uV 10:06:05 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 22.372 uVs 10:06:05 Integrated Voltage Measurement... 10:06:14 Voltage Offset Correction: v_off_ini = -62.283 uV, v_off_fin = -64.988 uV 10:06:14 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -29.249 uVs 10:06:14 Integrated Voltage Measurement... 10:06:23 Voltage Offset Correction: v_off_ini = -83.001 uV, v_off_fin = -89.663 uV 10:06:23 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 11.314 uVs 10:06:23 Integrated Voltage Measurement... 10:06:32 Voltage Offset Correction: v_off_ini = -70.294 uV, v_off_fin = -62.167 uV 10:06:32 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -19.972 uVs 10:06:32 Integrated Voltage Measurement... 10:06:41 Voltage Offset Correction: v_off_ini = -66.326 uV, v_off_fin = -81.383 uV 10:06:41 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 10.442 uVs 10:06:43 I2X Average and RMS of Measurements VT = -20.314 +- 4.183 uVs I2X = 1.869 +- 0.385 uTm^2 10:06:49 Wire Position: X0 = -2.0000 mm, Y0 = 0.0000 mm 10:06:51 Integrated Voltage Measurement... 10:07:00 Voltage Offset Correction: v_off_ini = -57.075 uV, v_off_fin = -80.923 uV 10:07:00 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -199.759 uVs 10:07:00 Integrated Voltage Measurement... 10:07:09 Voltage Offset Correction: v_off_ini = -73.971 uV, v_off_fin = -87.508 uV 10:07:09 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 186.838 uVs 10:07:09 Integrated Voltage Measurement... 10:07:18 Voltage Offset Correction: v_off_ini = -85.162 uV, v_off_fin = -81.223 uV 10:07:18 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -194.770 uVs 10:07:18 Integrated Voltage Measurement... 10:07:27 Voltage Offset Correction: v_off_ini = -66.760 uV, v_off_fin = -71.939 uV 10:07:27 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 177.107 uVs 10:07:27 Integrated Voltage Measurement... 10:07:36 Voltage Offset Correction: v_off_ini = -73.239 uV, v_off_fin = -79.868 uV 10:07:36 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -196.885 uVs 10:07:36 Integrated Voltage Measurement... 10:07:45 Voltage Offset Correction: v_off_ini = -79.318 uV, v_off_fin = -75.759 uV 10:07:45 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 186.966 uVs 10:07:48 I2Y Average and RMS of Measurements VT = -190.387 +- 3.195 uVs I2Y = -17.516 +- 0.294 uTm^2 10:07:52 Wire Position: X0 = -1.0000 mm, Y0 = 0.0000 mm 10:07:57 Integrated Voltage Measurement... 10:08:06 Voltage Offset Correction: v_off_ini = -86.254 uV, v_off_fin = -89.613 uV 10:08:06 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -63.799 uVs 10:08:06 Integrated Voltage Measurement... 10:08:15 Voltage Offset Correction: v_off_ini = -83.355 uV, v_off_fin = -71.812 uV 10:08:15 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 49.554 uVs 10:08:15 Integrated Voltage Measurement... 10:08:24 Voltage Offset Correction: v_off_ini = -80.988 uV, v_off_fin = -90.194 uV 10:08:24 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -54.148 uVs 10:08:24 Integrated Voltage Measurement... 10:08:33 Voltage Offset Correction: v_off_ini = -83.299 uV, v_off_fin = -87.288 uV 10:08:33 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 32.729 uVs 10:08:33 Integrated Voltage Measurement... 10:08:42 Voltage Offset Correction: v_off_ini = -95.496 uV, v_off_fin = -94.925 uV 10:08:42 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -49.225 uVs 10:08:42 Integrated Voltage Measurement... 10:08:51 Voltage Offset Correction: v_off_ini = -91.780 uV, v_off_fin = -76.019 uV 10:08:51 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 44.982 uVs 10:08:55 I1X Average and RMS of Measurements VT = -49.073 +- 5.581 uVs I1X = 9.815 +- 1.116 uTm 10:08:57 Wire Position: X0 = -1.0000 mm, Y0 = 0.0000 mm 10:09:02 Integrated Voltage Measurement... 10:09:11 Voltage Offset Correction: v_off_ini = -77.226 uV, v_off_fin = -88.486 uV 10:09:11 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -194.444 uVs 10:09:11 Integrated Voltage Measurement... 10:09:20 Voltage Offset Correction: v_off_ini = -85.724 uV, v_off_fin = -78.003 uV 10:09:20 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 182.668 uVs 10:09:20 Integrated Voltage Measurement... 10:09:29 Voltage Offset Correction: v_off_ini = -78.707 uV, v_off_fin = -89.170 uV 10:09:29 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -192.315 uVs 10:09:29 Integrated Voltage Measurement... 10:09:38 Voltage Offset Correction: v_off_ini = -81.628 uV, v_off_fin = -76.057 uV 10:09:38 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.4999 mm dXY = -0.4999 mm VT = 169.679 uVs 10:09:38 Integrated Voltage Measurement... 10:09:47 Voltage Offset Correction: v_off_ini = -82.090 uV, v_off_fin = -81.787 uV 10:09:47 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -193.822 uVs 10:09:47 Integrated Voltage Measurement... 10:09:56 Voltage Offset Correction: v_off_ini = -65.172 uV, v_off_fin = -71.486 uV 10:09:56 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 185.383 uVs 10:10:00 I1Y Average and RMS of Measurements VT = -186.385 +- 3.834 uVs I1Y = -37.278 +- 0.765 uTm 10:10:05 Wire Position: X0 = -1.0000 mm, Y0 = 0.0000 mm 10:10:08 Integrated Voltage Measurement... 10:10:17 Voltage Offset Correction: v_off_ini = -51.314 uV, v_off_fin = -69.123 uV 10:10:17 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -62.422 uVs 10:10:17 Integrated Voltage Measurement... 10:10:26 Voltage Offset Correction: v_off_ini = -69.521 uV, v_off_fin = -91.472 uV 10:10:26 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 29.356 uVs 10:10:26 Integrated Voltage Measurement... 10:10:35 Voltage Offset Correction: v_off_ini = -103.386 uV, v_off_fin = -74.874 uV 10:10:35 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -52.394 uVs 10:10:35 Integrated Voltage Measurement... 10:10:44 Voltage Offset Correction: v_off_ini = -97.108 uV, v_off_fin = -97.731 uV 10:10:44 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 47.397 uVs 10:10:44 Integrated Voltage Measurement... 10:10:53 Voltage Offset Correction: v_off_ini = -79.645 uV, v_off_fin = -85.190 uV 10:10:53 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -27.118 uVs 10:10:53 Integrated Voltage Measurement... 10:11:02 Voltage Offset Correction: v_off_ini = -82.934 uV, v_off_fin = -97.743 uV 10:11:02 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 44.732 uVs 10:11:04 I2X Average and RMS of Measurements VT = -43.903 +- 5.874 uVs I2X = 4.039 +- 0.540 uTm^2 10:11:09 Wire Position: X0 = -1.0000 mm, Y0 = 0.0000 mm 10:11:12 Integrated Voltage Measurement... 10:11:21 Voltage Offset Correction: v_off_ini = -89.232 uV, v_off_fin = -103.145 uV 10:11:21 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -188.666 uVs 10:11:21 Integrated Voltage Measurement... 10:11:30 Voltage Offset Correction: v_off_ini = -97.175 uV, v_off_fin = -99.774 uV 10:11:30 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 192.788 uVs 10:11:30 Integrated Voltage Measurement... 10:11:39 Voltage Offset Correction: v_off_ini = -93.682 uV, v_off_fin = -99.011 uV 10:11:39 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = -185.173 uVs 10:11:39 Integrated Voltage Measurement... 10:11:48 Voltage Offset Correction: v_off_ini = -85.934 uV, v_off_fin = -79.710 uV 10:11:48 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = 180.476 uVs 10:11:48 Integrated Voltage Measurement... 10:11:57 Voltage Offset Correction: v_off_ini = -84.424 uV, v_off_fin = -79.573 uV 10:11:57 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -183.394 uVs 10:11:57 Integrated Voltage Measurement... 10:12:06 Voltage Offset Correction: v_off_ini = -99.822 uV, v_off_fin = -95.002 uV 10:12:06 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 204.666 uVs 10:12:08 I2Y Average and RMS of Measurements VT = -189.194 +- 4.701 uVs I2Y = -17.406 +- 0.432 uTm^2 10:12:13 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:12:17 Integrated Voltage Measurement... 10:12:26 Voltage Offset Correction: v_off_ini = -82.969 uV, v_off_fin = -93.142 uV 10:12:26 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -75.756 uVs 10:12:26 Integrated Voltage Measurement... 10:12:35 Voltage Offset Correction: v_off_ini = -91.494 uV, v_off_fin = -91.971 uV 10:12:35 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 53.099 uVs 10:12:35 Integrated Voltage Measurement... 10:12:44 Voltage Offset Correction: v_off_ini = -96.052 uV, v_off_fin = -88.386 uV 10:12:44 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -69.065 uVs 10:12:44 Integrated Voltage Measurement... 10:12:53 Voltage Offset Correction: v_off_ini = -80.832 uV, v_off_fin = -98.218 uV 10:12:53 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 65.288 uVs 10:12:54 Integrated Voltage Measurement... 10:13:02 Voltage Offset Correction: v_off_ini = -94.349 uV, v_off_fin = -96.977 uV 10:13:02 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -50.180 uVs 10:13:03 Integrated Voltage Measurement... 10:13:11 Voltage Offset Correction: v_off_ini = -90.883 uV, v_off_fin = -91.301 uV 10:13:12 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 74.632 uVs 10:13:16 I1X Average and RMS of Measurements VT = -64.670 +- 1.955 uVs I1X = 12.934 +- 0.391 uTm 10:13:18 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:13:22 Integrated Voltage Measurement... 10:13:31 Voltage Offset Correction: v_off_ini = -84.830 uV, v_off_fin = -85.994 uV 10:13:31 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -209.430 uVs 10:13:31 Integrated Voltage Measurement... 10:13:40 Voltage Offset Correction: v_off_ini = -99.020 uV, v_off_fin = -70.067 uV 10:13:40 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 211.596 uVs 10:13:40 Integrated Voltage Measurement... 10:13:49 Voltage Offset Correction: v_off_ini = -99.417 uV, v_off_fin = -100.924 uV 10:13:49 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -188.710 uVs 10:13:49 Integrated Voltage Measurement... 10:13:58 Voltage Offset Correction: v_off_ini = -82.790 uV, v_off_fin = -76.070 uV 10:13:58 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.4999 mm dXY = -0.4999 mm VT = 209.967 uVs 10:13:58 Integrated Voltage Measurement... 10:14:07 Voltage Offset Correction: v_off_ini = -81.376 uV, v_off_fin = -99.949 uV 10:14:07 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -185.756 uVs 10:14:07 Integrated Voltage Measurement... 10:14:16 Voltage Offset Correction: v_off_ini = -78.582 uV, v_off_fin = -92.089 uV 10:14:16 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 185.118 uVs 10:14:21 I1Y Average and RMS of Measurements VT = -198.429 +- 10.257 uVs I1Y = -39.687 +- 2.052 uTm 10:14:26 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:14:28 Integrated Voltage Measurement... 10:14:37 Voltage Offset Correction: v_off_ini = -77.378 uV, v_off_fin = -82.857 uV 10:14:37 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -81.638 uVs 10:14:37 Integrated Voltage Measurement... 10:14:46 Voltage Offset Correction: v_off_ini = -86.151 uV, v_off_fin = -86.913 uV 10:14:46 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 72.248 uVs 10:14:46 Integrated Voltage Measurement... 10:14:55 Voltage Offset Correction: v_off_ini = -99.758 uV, v_off_fin = -110.436 uV 10:14:55 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -75.488 uVs 10:14:56 Integrated Voltage Measurement... 10:15:04 Voltage Offset Correction: v_off_ini = -106.054 uV, v_off_fin = -103.021 uV 10:15:05 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 83.441 uVs 10:15:05 Integrated Voltage Measurement... 10:15:14 Voltage Offset Correction: v_off_ini = -88.062 uV, v_off_fin = -77.406 uV 10:15:14 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -63.125 uVs 10:15:14 Integrated Voltage Measurement... 10:15:23 Voltage Offset Correction: v_off_ini = -69.556 uV, v_off_fin = -88.356 uV 10:15:23 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 46.611 uVs 10:15:25 I2X Average and RMS of Measurements VT = -70.425 +- 11.049 uVs I2X = 6.479 +- 1.016 uTm^2 10:15:30 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:15:33 Integrated Voltage Measurement... 10:15:42 Voltage Offset Correction: v_off_ini = -84.965 uV, v_off_fin = -79.562 uV 10:15:42 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -198.533 uVs 10:15:42 Integrated Voltage Measurement... 10:15:51 Voltage Offset Correction: v_off_ini = -92.065 uV, v_off_fin = -86.144 uV 10:15:51 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 220.682 uVs 10:15:51 Integrated Voltage Measurement... 10:16:00 Voltage Offset Correction: v_off_ini = -95.509 uV, v_off_fin = -91.681 uV 10:16:00 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -196.884 uVs 10:16:00 Integrated Voltage Measurement... 10:16:09 Voltage Offset Correction: v_off_ini = -91.117 uV, v_off_fin = -93.203 uV 10:16:09 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 215.164 uVs 10:16:09 Integrated Voltage Measurement... 10:16:18 Voltage Offset Correction: v_off_ini = -95.745 uV, v_off_fin = -82.404 uV 10:16:18 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -200.907 uVs 10:16:18 Integrated Voltage Measurement... 10:16:27 Voltage Offset Correction: v_off_ini = -71.457 uV, v_off_fin = -79.120 uV 10:16:27 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 190.172 uVs 10:16:29 I2Y Average and RMS of Measurements VT = -203.724 +- 5.969 uVs I2Y = -18.743 +- 0.549 uTm^2 10:16:34 Wire Position: X0 = 1.0000 mm, Y0 = 0.0000 mm 10:16:38 Integrated Voltage Measurement... 10:16:47 Voltage Offset Correction: v_off_ini = -101.684 uV, v_off_fin = -79.309 uV 10:16:47 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -49.043 uVs 10:16:47 Integrated Voltage Measurement... 10:16:56 Voltage Offset Correction: v_off_ini = -84.755 uV, v_off_fin = -77.656 uV 10:16:56 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 93.908 uVs 10:16:56 Integrated Voltage Measurement... 10:17:05 Voltage Offset Correction: v_off_ini = -75.347 uV, v_off_fin = -77.345 uV 10:17:05 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -78.911 uVs 10:17:05 Integrated Voltage Measurement... 10:17:14 Voltage Offset Correction: v_off_ini = -94.491 uV, v_off_fin = -84.759 uV 10:17:14 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 102.161 uVs 10:17:14 Integrated Voltage Measurement... 10:17:23 Voltage Offset Correction: v_off_ini = -82.477 uV, v_off_fin = -78.542 uV 10:17:23 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -85.139 uVs 10:17:23 Integrated Voltage Measurement... 10:17:32 Voltage Offset Correction: v_off_ini = -80.212 uV, v_off_fin = -78.834 uV 10:17:32 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 77.208 uVs 10:17:37 I1X Average and RMS of Measurements VT = -81.062 +- 7.782 uVs I1X = 16.212 +- 1.556 uTm 10:17:39 Wire Position: X0 = 1.0000 mm, Y0 = 0.0000 mm 10:17:43 Integrated Voltage Measurement... 10:17:52 Voltage Offset Correction: v_off_ini = -75.548 uV, v_off_fin = -85.531 uV 10:17:52 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -168.902 uVs 10:17:52 Integrated Voltage Measurement... 10:18:01 Voltage Offset Correction: v_off_ini = -89.409 uV, v_off_fin = -69.155 uV 10:18:01 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 203.822 uVs 10:18:01 Integrated Voltage Measurement... 10:18:10 Voltage Offset Correction: v_off_ini = -71.995 uV, v_off_fin = -81.038 uV 10:18:10 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -185.084 uVs 10:18:10 Integrated Voltage Measurement... 10:18:19 Voltage Offset Correction: v_off_ini = -80.784 uV, v_off_fin = -57.528 uV 10:18:19 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 196.278 uVs 10:18:19 Integrated Voltage Measurement... 10:18:28 Voltage Offset Correction: v_off_ini = -66.349 uV, v_off_fin = -85.730 uV 10:18:28 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -191.391 uVs 10:18:28 Integrated Voltage Measurement... 10:18:37 Voltage Offset Correction: v_off_ini = -72.607 uV, v_off_fin = -73.743 uV 10:18:37 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 181.517 uVs 10:18:41 I1Y Average and RMS of Measurements VT = -187.832 +- 2.015 uVs I1Y = -37.566 +- 0.403 uTm 10:18:47 Wire Position: X0 = 1.0000 mm, Y0 = 0.0000 mm 10:18:49 Integrated Voltage Measurement... 10:18:58 Voltage Offset Correction: v_off_ini = -83.226 uV, v_off_fin = -80.129 uV 10:18:58 Integrated Voltage Measurement dXYus = -1.0001 mm, dXYds = 0.0000 mm dXY = -1.0001 mm VT = -72.398 uVs 10:18:58 Integrated Voltage Measurement... 10:19:07 Voltage Offset Correction: v_off_ini = -87.430 uV, v_off_fin = -101.779 uV 10:19:07 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 107.976 uVs 10:19:07 Integrated Voltage Measurement... 10:19:16 Voltage Offset Correction: v_off_ini = -104.477 uV, v_off_fin = -85.589 uV 10:19:16 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -64.551 uVs 10:19:16 Integrated Voltage Measurement... 10:19:25 Voltage Offset Correction: v_off_ini = -76.647 uV, v_off_fin = -101.123 uV 10:19:25 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 65.763 uVs 10:19:25 Integrated Voltage Measurement... 10:19:34 Voltage Offset Correction: v_off_ini = -80.378 uV, v_off_fin = -79.590 uV 10:19:34 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -94.220 uVs 10:19:34 Integrated Voltage Measurement... 10:19:43 Voltage Offset Correction: v_off_ini = -99.791 uV, v_off_fin = -113.407 uV 10:19:43 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 93.808 uVs 10:19:46 I2X Average and RMS of Measurements VT = -83.119 +- 12.797 uVs I2X = 7.647 +- 1.177 uTm^2 10:19:51 Wire Position: X0 = 1.0000 mm, Y0 = 0.0000 mm 10:19:53 Integrated Voltage Measurement... 10:20:02 Voltage Offset Correction: v_off_ini = -89.006 uV, v_off_fin = -90.379 uV 10:20:03 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -187.276 uVs 10:20:03 Integrated Voltage Measurement... 10:20:12 Voltage Offset Correction: v_off_ini = -92.243 uV, v_off_fin = -93.937 uV 10:20:12 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 205.784 uVs 10:20:12 Integrated Voltage Measurement... 10:20:21 Voltage Offset Correction: v_off_ini = -84.638 uV, v_off_fin = -88.045 uV 10:20:21 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -205.931 uVs 10:20:21 Integrated Voltage Measurement... 10:20:30 Voltage Offset Correction: v_off_ini = -87.567 uV, v_off_fin = -94.314 uV 10:20:30 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 176.948 uVs 10:20:30 Integrated Voltage Measurement... 10:20:39 Voltage Offset Correction: v_off_ini = -101.106 uV, v_off_fin = -93.970 uV 10:20:39 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -220.641 uVs 10:20:39 Integrated Voltage Measurement... 10:20:48 Voltage Offset Correction: v_off_ini = -96.489 uV, v_off_fin = -80.352 uV 10:20:48 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = 196.400 uVs 10:20:50 I2Y Average and RMS of Measurements VT = -198.830 +- 7.160 uVs I2Y = -18.293 +- 0.659 uTm^2 10:20:55 Wire Position: X0 = 2.0000 mm, Y0 = 0.0000 mm 10:20:59 Integrated Voltage Measurement... 10:21:08 Voltage Offset Correction: v_off_ini = -92.615 uV, v_off_fin = -95.192 uV 10:21:08 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -75.403 uVs 10:21:08 Integrated Voltage Measurement... 10:21:17 Voltage Offset Correction: v_off_ini = -83.528 uV, v_off_fin = -89.068 uV 10:21:17 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 76.433 uVs 10:21:17 Integrated Voltage Measurement... 10:21:26 Voltage Offset Correction: v_off_ini = -102.965 uV, v_off_fin = -117.219 uV 10:21:26 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -97.872 uVs 10:21:26 Integrated Voltage Measurement... 10:21:35 Voltage Offset Correction: v_off_ini = -124.339 uV, v_off_fin = -106.292 uV 10:21:35 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 109.074 uVs 10:21:35 Integrated Voltage Measurement... 10:21:44 Voltage Offset Correction: v_off_ini = -101.333 uV, v_off_fin = -106.734 uV 10:21:44 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -114.116 uVs 10:21:44 Integrated Voltage Measurement... 10:21:53 Voltage Offset Correction: v_off_ini = -114.819 uV, v_off_fin = -120.643 uV 10:21:53 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 72.747 uVs 10:21:57 I1X Average and RMS of Measurements VT = -90.941 +- 11.386 uVs I1X = 18.188 +- 2.277 uTm 10:22:00 Wire Position: X0 = 2.0000 mm, Y0 = 0.0000 mm 10:22:04 Integrated Voltage Measurement... 10:22:13 Voltage Offset Correction: v_off_ini = -113.161 uV, v_off_fin = -119.504 uV 10:22:13 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -191.189 uVs 10:22:13 Integrated Voltage Measurement... 10:22:22 Voltage Offset Correction: v_off_ini = -116.469 uV, v_off_fin = -106.468 uV 10:22:22 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 203.544 uVs 10:22:22 Integrated Voltage Measurement... 10:22:31 Voltage Offset Correction: v_off_ini = -103.933 uV, v_off_fin = -104.072 uV 10:22:31 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -200.945 uVs 10:22:31 Integrated Voltage Measurement... 10:22:40 Voltage Offset Correction: v_off_ini = -111.940 uV, v_off_fin = -91.070 uV 10:22:40 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 189.149 uVs 10:22:40 Integrated Voltage Measurement... 10:22:49 Voltage Offset Correction: v_off_ini = -102.350 uV, v_off_fin = -106.389 uV 10:22:49 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -210.070 uVs 10:22:49 Integrated Voltage Measurement... 10:22:58 Voltage Offset Correction: v_off_ini = -100.111 uV, v_off_fin = -98.955 uV 10:22:58 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 213.363 uVs 10:23:02 I1Y Average and RMS of Measurements VT = -201.377 +- 7.372 uVs I1Y = -40.275 +- 1.474 uTm 10:23:07 Wire Position: X0 = 2.0000 mm, Y0 = 0.0000 mm 10:23:10 Integrated Voltage Measurement... 10:23:19 Voltage Offset Correction: v_off_ini = -108.968 uV, v_off_fin = -92.332 uV 10:23:19 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -103.724 uVs 10:23:19 Integrated Voltage Measurement... 10:23:28 Voltage Offset Correction: v_off_ini = -94.214 uV, v_off_fin = -103.527 uV 10:23:28 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 108.004 uVs 10:23:28 Integrated Voltage Measurement... 10:23:37 Voltage Offset Correction: v_off_ini = -97.237 uV, v_off_fin = -99.571 uV 10:23:37 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -99.660 uVs 10:23:37 Integrated Voltage Measurement... 10:23:46 Voltage Offset Correction: v_off_ini = -104.887 uV, v_off_fin = -121.584 uV 10:23:46 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 72.701 uVs 10:23:46 Integrated Voltage Measurement... 10:23:55 Voltage Offset Correction: v_off_ini = -98.571 uV, v_off_fin = -82.446 uV 10:23:55 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -94.620 uVs 10:23:55 Integrated Voltage Measurement... 10:24:04 Voltage Offset Correction: v_off_ini = -89.448 uV, v_off_fin = -100.250 uV 10:24:04 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 89.417 uVs 10:24:07 I2X Average and RMS of Measurements VT = -94.688 +- 8.254 uVs I2X = 8.711 +- 0.759 uTm^2 10:24:12 Wire Position: X0 = 2.0000 mm, Y0 = 0.0000 mm 10:24:14 Integrated Voltage Measurement... 10:24:23 Voltage Offset Correction: v_off_ini = -100.106 uV, v_off_fin = -105.744 uV 10:24:23 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -208.895 uVs 10:24:23 Integrated Voltage Measurement... 10:24:32 Voltage Offset Correction: v_off_ini = -100.038 uV, v_off_fin = -92.458 uV 10:24:32 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 215.050 uVs 10:24:32 Integrated Voltage Measurement... 10:24:41 Voltage Offset Correction: v_off_ini = -89.198 uV, v_off_fin = -84.215 uV 10:24:41 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -204.956 uVs 10:24:41 Integrated Voltage Measurement... 10:24:50 Voltage Offset Correction: v_off_ini = -108.297 uV, v_off_fin = -101.790 uV 10:24:50 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 206.946 uVs 10:24:50 Integrated Voltage Measurement... 10:24:59 Voltage Offset Correction: v_off_ini = -111.196 uV, v_off_fin = -87.333 uV 10:24:59 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -206.867 uVs 10:25:00 Integrated Voltage Measurement... 10:25:08 Voltage Offset Correction: v_off_ini = -105.560 uV, v_off_fin = -94.432 uV 10:25:09 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 234.880 uVs 10:25:11 I2Y Average and RMS of Measurements VT = -212.932 +- 6.130 uVs I2Y = -19.590 +- 0.564 uTm^2 10:25:19 Wire Position: X0 = 0.0000 mm, Y0 = -1.0000 mm 10:25:23 Integrated Voltage Measurement... 10:25:32 Voltage Offset Correction: v_off_ini = -113.952 uV, v_off_fin = -93.024 uV 10:25:32 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -62.793 uVs 10:25:32 Integrated Voltage Measurement... 10:25:41 Voltage Offset Correction: v_off_ini = -89.116 uV, v_off_fin = -82.991 uV 10:25:41 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 48.727 uVs 10:25:41 Integrated Voltage Measurement... 10:25:50 Voltage Offset Correction: v_off_ini = -78.502 uV, v_off_fin = -87.760 uV 10:25:50 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -55.575 uVs 10:25:50 Integrated Voltage Measurement... 10:25:59 Voltage Offset Correction: v_off_ini = -77.604 uV, v_off_fin = -86.431 uV 10:25:59 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 40.189 uVs 10:25:59 Integrated Voltage Measurement... 10:26:08 Voltage Offset Correction: v_off_ini = -86.207 uV, v_off_fin = -91.738 uV 10:26:08 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -59.883 uVs 10:26:08 Integrated Voltage Measurement... 10:26:17 Voltage Offset Correction: v_off_ini = -110.352 uV, v_off_fin = -101.362 uV 10:26:17 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 62.131 uVs 10:26:21 I1X Average and RMS of Measurements VT = -54.883 +- 5.394 uVs I1X = 10.977 +- 1.079 uTm 10:26:24 Wire Position: X0 = 0.0000 mm, Y0 = -1.0000 mm 10:26:28 Integrated Voltage Measurement... 10:26:37 Voltage Offset Correction: v_off_ini = -91.107 uV, v_off_fin = -106.613 uV 10:26:37 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -172.261 uVs 10:26:37 Integrated Voltage Measurement... 10:26:46 Voltage Offset Correction: v_off_ini = -85.890 uV, v_off_fin = -84.003 uV 10:26:46 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 186.977 uVs 10:26:46 Integrated Voltage Measurement... 10:26:55 Voltage Offset Correction: v_off_ini = -93.749 uV, v_off_fin = -108.814 uV 10:26:55 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -171.094 uVs 10:26:55 Integrated Voltage Measurement... 10:27:04 Voltage Offset Correction: v_off_ini = -83.888 uV, v_off_fin = -82.463 uV 10:27:04 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 185.759 uVs 10:27:04 Integrated Voltage Measurement... 10:27:13 Voltage Offset Correction: v_off_ini = -107.782 uV, v_off_fin = -107.778 uV 10:27:13 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -185.195 uVs 10:27:13 Integrated Voltage Measurement... 10:27:22 Voltage Offset Correction: v_off_ini = -104.923 uV, v_off_fin = -83.696 uV 10:27:22 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 186.388 uVs 10:27:26 I1Y Average and RMS of Measurements VT = -181.279 +- 3.228 uVs I1Y = -36.256 +- 0.646 uTm 10:27:29 Wire Position: X0 = 0.0000 mm, Y0 = -1.0000 mm 10:27:31 Integrated Voltage Measurement... 10:27:40 Voltage Offset Correction: v_off_ini = -76.002 uV, v_off_fin = -96.477 uV 10:27:40 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -77.783 uVs 10:27:40 Integrated Voltage Measurement... 10:27:49 Voltage Offset Correction: v_off_ini = -90.288 uV, v_off_fin = -106.523 uV 10:27:49 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 48.852 uVs 10:27:49 Integrated Voltage Measurement... 10:27:58 Voltage Offset Correction: v_off_ini = -83.423 uV, v_off_fin = -88.863 uV 10:27:58 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -67.873 uVs 10:27:58 Integrated Voltage Measurement... 10:28:07 Voltage Offset Correction: v_off_ini = -106.768 uV, v_off_fin = -94.688 uV 10:28:07 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 39.592 uVs 10:28:07 Integrated Voltage Measurement... 10:28:16 Voltage Offset Correction: v_off_ini = -82.367 uV, v_off_fin = -84.275 uV 10:28:16 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -63.168 uVs 10:28:16 Integrated Voltage Measurement... 10:28:25 Voltage Offset Correction: v_off_ini = -84.519 uV, v_off_fin = -101.745 uV 10:28:25 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 54.974 uVs 10:28:28 I2X Average and RMS of Measurements VT = -58.707 +- 3.922 uVs I2X = 5.401 +- 0.361 uTm^2 10:28:30 Wire Position: X0 = 0.0000 mm, Y0 = -1.0000 mm 10:28:33 Integrated Voltage Measurement... 10:28:42 Voltage Offset Correction: v_off_ini = -95.080 uV, v_off_fin = -89.176 uV 10:28:42 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -163.154 uVs 10:28:42 Integrated Voltage Measurement... 10:28:51 Voltage Offset Correction: v_off_ini = -81.092 uV, v_off_fin = -106.076 uV 10:28:51 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 163.487 uVs 10:28:51 Integrated Voltage Measurement... 10:29:00 Voltage Offset Correction: v_off_ini = -107.208 uV, v_off_fin = -90.671 uV 10:29:00 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -174.031 uVs 10:29:00 Integrated Voltage Measurement... 10:29:09 Voltage Offset Correction: v_off_ini = -101.000 uV, v_off_fin = -91.355 uV 10:29:09 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 186.187 uVs 10:29:09 Integrated Voltage Measurement... 10:29:18 Voltage Offset Correction: v_off_ini = -93.431 uV, v_off_fin = -106.215 uV 10:29:18 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -174.221 uVs 10:29:18 Integrated Voltage Measurement... 10:29:27 Voltage Offset Correction: v_off_ini = -94.807 uV, v_off_fin = -90.830 uV 10:29:27 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 185.154 uVs 10:29:30 I2Y Average and RMS of Measurements VT = -174.372 +- 7.817 uVs I2Y = -16.042 +- 0.719 uTm^2 10:29:34 Wire Position: X0 = 0.0000 mm, Y0 = -0.5000 mm 10:29:38 Integrated Voltage Measurement... 10:29:47 Voltage Offset Correction: v_off_ini = -112.182 uV, v_off_fin = -125.262 uV 10:29:47 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -54.239 uVs 10:29:47 Integrated Voltage Measurement... 10:29:56 Voltage Offset Correction: v_off_ini = -103.511 uV, v_off_fin = -101.334 uV 10:29:56 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 66.887 uVs 10:29:56 Integrated Voltage Measurement... 10:30:05 Voltage Offset Correction: v_off_ini = -88.860 uV, v_off_fin = -106.284 uV 10:30:05 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -70.343 uVs 10:30:05 Integrated Voltage Measurement... 10:30:14 Voltage Offset Correction: v_off_ini = -94.297 uV, v_off_fin = -93.287 uV 10:30:14 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 53.744 uVs 10:30:14 Integrated Voltage Measurement... 10:30:23 Voltage Offset Correction: v_off_ini = -92.007 uV, v_off_fin = -108.403 uV 10:30:23 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -55.666 uVs 10:30:23 Integrated Voltage Measurement... 10:30:32 Voltage Offset Correction: v_off_ini = -97.813 uV, v_off_fin = -88.661 uV 10:30:32 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 61.844 uVs 10:30:36 I1X Average and RMS of Measurements VT = -60.454 +- 1.344 uVs I1X = 12.091 +- 0.269 uTm 10:30:39 Wire Position: X0 = 0.0000 mm, Y0 = -0.5000 mm 10:30:43 Integrated Voltage Measurement... 10:30:52 Voltage Offset Correction: v_off_ini = -111.330 uV, v_off_fin = -113.243 uV 10:30:52 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -202.406 uVs 10:30:52 Integrated Voltage Measurement... 10:31:01 Voltage Offset Correction: v_off_ini = -102.409 uV, v_off_fin = -93.924 uV 10:31:01 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 175.717 uVs 10:31:01 Integrated Voltage Measurement... 10:31:10 Voltage Offset Correction: v_off_ini = -96.933 uV, v_off_fin = -105.747 uV 10:31:10 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -198.053 uVs 10:31:10 Integrated Voltage Measurement... 10:31:19 Voltage Offset Correction: v_off_ini = -103.774 uV, v_off_fin = -100.332 uV 10:31:19 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 181.429 uVs 10:31:19 Integrated Voltage Measurement... 10:31:28 Voltage Offset Correction: v_off_ini = -109.601 uV, v_off_fin = -115.433 uV 10:31:28 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -194.520 uVs 10:31:28 Integrated Voltage Measurement... 10:31:37 Voltage Offset Correction: v_off_ini = -99.833 uV, v_off_fin = -89.657 uV 10:31:37 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 186.218 uVs 10:31:41 I1Y Average and RMS of Measurements VT = -189.724 +- 0.534 uVs I1Y = -37.945 +- 0.107 uTm 10:31:44 Wire Position: X0 = 0.0000 mm, Y0 = -0.5000 mm 10:31:46 Integrated Voltage Measurement... 10:31:55 Voltage Offset Correction: v_off_ini = -93.505 uV, v_off_fin = -89.598 uV 10:31:55 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -68.205 uVs 10:31:55 Integrated Voltage Measurement... 10:32:04 Voltage Offset Correction: v_off_ini = -101.376 uV, v_off_fin = -109.780 uV 10:32:04 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 45.155 uVs 10:32:04 Integrated Voltage Measurement... 10:32:13 Voltage Offset Correction: v_off_ini = -90.825 uV, v_off_fin = -94.630 uV 10:32:13 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -51.841 uVs 10:32:13 Integrated Voltage Measurement... 10:32:22 Voltage Offset Correction: v_off_ini = -91.175 uV, v_off_fin = -109.709 uV 10:32:22 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 80.065 uVs 10:32:22 Integrated Voltage Measurement... 10:32:31 Voltage Offset Correction: v_off_ini = -92.077 uV, v_off_fin = -88.450 uV 10:32:31 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -46.024 uVs 10:32:31 Integrated Voltage Measurement... 10:32:40 Voltage Offset Correction: v_off_ini = -89.447 uV, v_off_fin = -100.034 uV 10:32:40 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 73.012 uVs 10:32:43 I2X Average and RMS of Measurements VT = -60.717 +- 3.879 uVs I2X = 5.586 +- 0.357 uTm^2 10:32:45 Wire Position: X0 = 0.0000 mm, Y0 = -0.5000 mm 10:32:48 Integrated Voltage Measurement... 10:32:57 Voltage Offset Correction: v_off_ini = -115.818 uV, v_off_fin = -111.917 uV 10:32:57 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -213.561 uVs 10:32:57 Integrated Voltage Measurement... 10:33:06 Voltage Offset Correction: v_off_ini = -127.247 uV, v_off_fin = -119.971 uV 10:33:06 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 202.576 uVs 10:33:06 Integrated Voltage Measurement... 10:33:15 Voltage Offset Correction: v_off_ini = -112.377 uV, v_off_fin = -116.229 uV 10:33:15 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -204.708 uVs 10:33:15 Integrated Voltage Measurement... 10:33:24 Voltage Offset Correction: v_off_ini = -126.803 uV, v_off_fin = -114.493 uV 10:33:24 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 201.930 uVs 10:33:24 Integrated Voltage Measurement... 10:33:33 Voltage Offset Correction: v_off_ini = -114.192 uV, v_off_fin = -119.742 uV 10:33:33 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -196.988 uVs 10:33:33 Integrated Voltage Measurement... 10:33:42 Voltage Offset Correction: v_off_ini = -125.599 uV, v_off_fin = -118.920 uV 10:33:42 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 205.569 uVs 10:33:45 I2Y Average and RMS of Measurements VT = -204.222 +- 2.844 uVs I2Y = -18.788 +- 0.262 uTm^2 10:33:49 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:33:53 Integrated Voltage Measurement... 10:34:02 Voltage Offset Correction: v_off_ini = -137.018 uV, v_off_fin = -128.650 uV 10:34:02 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.4998 mm dXY = -0.4999 mm VT = -62.163 uVs 10:34:02 Integrated Voltage Measurement... 10:34:11 Voltage Offset Correction: v_off_ini = -119.128 uV, v_off_fin = -115.614 uV 10:34:11 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = 43.925 uVs 10:34:11 Integrated Voltage Measurement... 10:34:20 Voltage Offset Correction: v_off_ini = -124.565 uV, v_off_fin = -111.753 uV 10:34:20 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -57.857 uVs 10:34:20 Integrated Voltage Measurement... 10:34:29 Voltage Offset Correction: v_off_ini = -116.135 uV, v_off_fin = -123.260 uV 10:34:29 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 67.017 uVs 10:34:29 Integrated Voltage Measurement... 10:34:38 Voltage Offset Correction: v_off_ini = -124.639 uV, v_off_fin = -121.581 uV 10:34:38 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -52.567 uVs 10:34:38 Integrated Voltage Measurement... 10:34:47 Voltage Offset Correction: v_off_ini = -114.123 uV, v_off_fin = -103.629 uV 10:34:47 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 43.762 uVs 10:34:51 I1X Average and RMS of Measurements VT = -54.549 +- 5.923 uVs I1X = 10.911 +- 1.184 uTm 10:34:54 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:34:58 Integrated Voltage Measurement... 10:35:07 Voltage Offset Correction: v_off_ini = -133.904 uV, v_off_fin = -128.875 uV 10:35:07 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -193.189 uVs 10:35:07 Integrated Voltage Measurement... 10:35:16 Voltage Offset Correction: v_off_ini = -128.834 uV, v_off_fin = -109.084 uV 10:35:16 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 199.146 uVs 10:35:16 Integrated Voltage Measurement... 10:35:25 Voltage Offset Correction: v_off_ini = -121.406 uV, v_off_fin = -126.986 uV 10:35:25 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -194.491 uVs 10:35:25 Integrated Voltage Measurement... 10:35:34 Voltage Offset Correction: v_off_ini = -110.402 uV, v_off_fin = -110.926 uV 10:35:34 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 197.067 uVs 10:35:34 Integrated Voltage Measurement... 10:35:43 Voltage Offset Correction: v_off_ini = -110.491 uV, v_off_fin = -123.218 uV 10:35:43 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -204.834 uVs 10:35:43 Integrated Voltage Measurement... 10:35:52 Voltage Offset Correction: v_off_ini = -101.789 uV, v_off_fin = -90.064 uV 10:35:52 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 197.156 uVs 10:35:56 I1Y Average and RMS of Measurements VT = -197.647 +- 2.373 uVs I1Y = -39.529 +- 0.475 uTm 10:35:59 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:36:01 Integrated Voltage Measurement... 10:36:10 Voltage Offset Correction: v_off_ini = -96.391 uV, v_off_fin = -101.931 uV 10:36:10 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -71.022 uVs 10:36:10 Integrated Voltage Measurement... 10:36:19 Voltage Offset Correction: v_off_ini = -108.037 uV, v_off_fin = -99.763 uV 10:36:19 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 62.708 uVs 10:36:19 Integrated Voltage Measurement... 10:36:28 Voltage Offset Correction: v_off_ini = -94.157 uV, v_off_fin = -80.324 uV 10:36:28 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -75.188 uVs 10:36:29 Integrated Voltage Measurement... 10:36:38 Voltage Offset Correction: v_off_ini = -99.998 uV, v_off_fin = -112.233 uV 10:36:38 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 73.600 uVs 10:36:38 Integrated Voltage Measurement... 10:36:47 Voltage Offset Correction: v_off_ini = -114.602 uV, v_off_fin = -92.704 uV 10:36:47 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -46.249 uVs 10:36:47 Integrated Voltage Measurement... 10:36:56 Voltage Offset Correction: v_off_ini = -101.822 uV, v_off_fin = -105.491 uV 10:36:56 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 81.938 uVs 10:36:58 I2X Average and RMS of Measurements VT = -68.451 +- 4.352 uVs I2X = 6.298 +- 0.400 uTm^2 10:37:01 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:37:03 Integrated Voltage Measurement... 10:37:12 Voltage Offset Correction: v_off_ini = -97.651 uV, v_off_fin = -99.833 uV 10:37:12 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -201.446 uVs 10:37:12 Integrated Voltage Measurement... 10:37:21 Voltage Offset Correction: v_off_ini = -85.954 uV, v_off_fin = -90.167 uV 10:37:21 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 191.912 uVs 10:37:21 Integrated Voltage Measurement... 10:37:30 Voltage Offset Correction: v_off_ini = -98.745 uV, v_off_fin = -92.116 uV 10:37:30 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = -214.766 uVs 10:37:30 Integrated Voltage Measurement... 10:37:39 Voltage Offset Correction: v_off_ini = -96.879 uV, v_off_fin = -100.376 uV 10:37:39 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = 216.343 uVs 10:37:39 Integrated Voltage Measurement... 10:37:48 Voltage Offset Correction: v_off_ini = -98.301 uV, v_off_fin = -85.679 uV 10:37:48 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -203.978 uVs 10:37:48 Integrated Voltage Measurement... 10:37:57 Voltage Offset Correction: v_off_ini = -83.930 uV, v_off_fin = -72.300 uV 10:37:57 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 178.856 uVs 10:38:00 I2Y Average and RMS of Measurements VT = -201.217 +- 10.363 uVs I2Y = -18.513 +- 0.954 uTm^2 10:38:04 Wire Position: X0 = 0.0000 mm, Y0 = 0.5000 mm 10:38:08 Integrated Voltage Measurement... 10:38:17 Voltage Offset Correction: v_off_ini = -73.507 uV, v_off_fin = -81.479 uV 10:38:17 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -61.997 uVs 10:38:17 Integrated Voltage Measurement... 10:38:26 Voltage Offset Correction: v_off_ini = -82.669 uV, v_off_fin = -74.651 uV 10:38:26 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 61.924 uVs 10:38:26 Integrated Voltage Measurement... 10:38:35 Voltage Offset Correction: v_off_ini = -80.718 uV, v_off_fin = -75.862 uV 10:38:35 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -50.875 uVs 10:38:35 Integrated Voltage Measurement... 10:38:44 Voltage Offset Correction: v_off_ini = -76.306 uV, v_off_fin = -65.645 uV 10:38:44 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 50.977 uVs 10:38:44 Integrated Voltage Measurement... 10:38:53 Voltage Offset Correction: v_off_ini = -84.331 uV, v_off_fin = -88.431 uV 10:38:53 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -34.918 uVs 10:38:53 Integrated Voltage Measurement... 10:39:02 Voltage Offset Correction: v_off_ini = -86.147 uV, v_off_fin = -73.854 uV 10:39:02 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 44.322 uVs 10:39:06 I1X Average and RMS of Measurements VT = -50.835 +- 9.121 uVs I1X = 10.167 +- 1.824 uTm 10:39:09 Wire Position: X0 = 0.0000 mm, Y0 = 0.5000 mm 10:39:13 Integrated Voltage Measurement... 10:39:22 Voltage Offset Correction: v_off_ini = -91.649 uV, v_off_fin = -85.322 uV 10:39:22 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = -205.341 uVs 10:39:22 Integrated Voltage Measurement... 10:39:31 Voltage Offset Correction: v_off_ini = -80.512 uV, v_off_fin = -61.823 uV 10:39:31 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 205.715 uVs 10:39:31 Integrated Voltage Measurement... 10:39:40 Voltage Offset Correction: v_off_ini = -59.924 uV, v_off_fin = -65.318 uV 10:39:40 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -215.264 uVs 10:39:40 Integrated Voltage Measurement... 10:39:49 Voltage Offset Correction: v_off_ini = -80.221 uV, v_off_fin = -60.500 uV 10:39:49 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 192.851 uVs 10:39:49 Integrated Voltage Measurement... 10:39:58 Voltage Offset Correction: v_off_ini = -72.983 uV, v_off_fin = -93.583 uV 10:39:58 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -204.126 uVs 10:39:58 Integrated Voltage Measurement... 10:40:07 Voltage Offset Correction: v_off_ini = -73.578 uV, v_off_fin = -59.592 uV 10:40:07 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 198.107 uVs 10:40:11 I1Y Average and RMS of Measurements VT = -203.567 +- 1.834 uVs I1Y = -40.713 +- 0.368 uTm 10:40:14 Wire Position: X0 = 0.0000 mm, Y0 = 0.5000 mm 10:40:17 Integrated Voltage Measurement... 10:40:26 Voltage Offset Correction: v_off_ini = -56.277 uV, v_off_fin = -59.360 uV 10:40:26 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -50.537 uVs 10:40:26 Integrated Voltage Measurement... 10:40:35 Voltage Offset Correction: v_off_ini = -65.010 uV, v_off_fin = -60.743 uV 10:40:35 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 63.141 uVs 10:40:35 Integrated Voltage Measurement... 10:40:44 Voltage Offset Correction: v_off_ini = -49.318 uV, v_off_fin = -48.663 uV 10:40:44 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -48.555 uVs 10:40:44 Integrated Voltage Measurement... 10:40:53 Voltage Offset Correction: v_off_ini = -57.695 uV, v_off_fin = -51.037 uV 10:40:53 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 60.273 uVs 10:40:53 Integrated Voltage Measurement... 10:41:02 Voltage Offset Correction: v_off_ini = -38.979 uV, v_off_fin = -35.712 uV 10:41:02 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -71.943 uVs 10:41:02 Integrated Voltage Measurement... 10:41:11 Voltage Offset Correction: v_off_ini = -63.504 uV, v_off_fin = -71.253 uV 10:41:11 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 63.430 uVs 10:41:13 I2X Average and RMS of Measurements VT = -59.647 +- 5.771 uVs I2X = 5.488 +- 0.531 uTm^2 10:41:16 Wire Position: X0 = 0.0000 mm, Y0 = 0.5000 mm 10:41:18 Integrated Voltage Measurement... 10:41:27 Voltage Offset Correction: v_off_ini = -73.377 uV, v_off_fin = -73.871 uV 10:41:27 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -183.687 uVs 10:41:27 Integrated Voltage Measurement... 10:41:36 Voltage Offset Correction: v_off_ini = -68.896 uV, v_off_fin = -73.619 uV 10:41:36 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 215.587 uVs 10:41:36 Integrated Voltage Measurement... 10:41:45 Voltage Offset Correction: v_off_ini = -76.792 uV, v_off_fin = -62.960 uV 10:41:45 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -210.615 uVs 10:41:45 Integrated Voltage Measurement... 10:41:54 Voltage Offset Correction: v_off_ini = -62.555 uV, v_off_fin = -62.047 uV 10:41:55 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 208.636 uVs 10:41:55 Integrated Voltage Measurement... 10:42:04 Voltage Offset Correction: v_off_ini = -67.302 uV, v_off_fin = -69.846 uV 10:42:04 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -205.262 uVs 10:42:04 Integrated Voltage Measurement... 10:42:13 Voltage Offset Correction: v_off_ini = -75.098 uV, v_off_fin = -67.055 uV 10:42:13 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 227.004 uVs 10:42:15 I2Y Average and RMS of Measurements VT = -208.465 +- 6.784 uVs I2Y = -19.179 +- 0.624 uTm^2 10:42:19 Wire Position: X0 = 0.0000 mm, Y0 = 1.0000 mm 10:42:23 Integrated Voltage Measurement... 10:42:32 Voltage Offset Correction: v_off_ini = -86.636 uV, v_off_fin = -78.821 uV 10:42:32 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -26.784 uVs 10:42:32 Integrated Voltage Measurement... 10:42:41 Voltage Offset Correction: v_off_ini = -91.674 uV, v_off_fin = -81.493 uV 10:42:41 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 53.251 uVs 10:42:41 Integrated Voltage Measurement... 10:42:50 Voltage Offset Correction: v_off_ini = -79.771 uV, v_off_fin = -81.710 uV 10:42:50 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -54.718 uVs 10:42:50 Integrated Voltage Measurement... 10:42:59 Voltage Offset Correction: v_off_ini = -78.484 uV, v_off_fin = -76.083 uV 10:42:59 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 57.161 uVs 10:42:59 Integrated Voltage Measurement... 10:43:08 Voltage Offset Correction: v_off_ini = -84.013 uV, v_off_fin = -72.262 uV 10:43:08 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -56.249 uVs 10:43:08 Integrated Voltage Measurement... 10:43:17 Voltage Offset Correction: v_off_ini = -88.733 uV, v_off_fin = -75.804 uV 10:43:17 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 66.893 uVs 10:43:22 I1X Average and RMS of Measurements VT = -52.509 +- 9.127 uVs I1X = 10.502 +- 1.825 uTm 10:43:24 Wire Position: X0 = 0.0000 mm, Y0 = 1.0000 mm 10:43:28 Integrated Voltage Measurement... 10:43:37 Voltage Offset Correction: v_off_ini = -88.030 uV, v_off_fin = -86.255 uV 10:43:37 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -217.189 uVs 10:43:37 Integrated Voltage Measurement... 10:43:46 Voltage Offset Correction: v_off_ini = -99.239 uV, v_off_fin = -69.513 uV 10:43:46 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 212.059 uVs 10:43:46 Integrated Voltage Measurement... 10:43:55 Voltage Offset Correction: v_off_ini = -84.573 uV, v_off_fin = -98.617 uV 10:43:55 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -191.862 uVs 10:43:55 Integrated Voltage Measurement... 10:44:04 Voltage Offset Correction: v_off_ini = -92.548 uV, v_off_fin = -83.808 uV 10:44:04 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 248.898 uVs 10:44:04 Integrated Voltage Measurement... 10:44:13 Voltage Offset Correction: v_off_ini = -90.176 uV, v_off_fin = -95.492 uV 10:44:13 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -221.822 uVs 10:44:13 Integrated Voltage Measurement... 10:44:22 Voltage Offset Correction: v_off_ini = -87.454 uV, v_off_fin = -83.338 uV 10:44:22 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 209.264 uVs 10:44:26 I1Y Average and RMS of Measurements VT = -216.849 +- 2.525 uVs I1Y = -43.370 +- 0.505 uTm 10:44:29 Wire Position: X0 = 0.0000 mm, Y0 = 1.0000 mm 10:44:31 Integrated Voltage Measurement... 10:44:40 Voltage Offset Correction: v_off_ini = -82.908 uV, v_off_fin = -78.750 uV 10:44:40 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -63.203 uVs 10:44:41 Integrated Voltage Measurement... 10:44:49 Voltage Offset Correction: v_off_ini = -92.327 uV, v_off_fin = -99.021 uV 10:44:50 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 69.416 uVs 10:44:50 Integrated Voltage Measurement... 10:44:58 Voltage Offset Correction: v_off_ini = -95.752 uV, v_off_fin = -88.871 uV 10:44:59 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -39.621 uVs 10:44:59 Integrated Voltage Measurement... 10:45:08 Voltage Offset Correction: v_off_ini = -86.759 uV, v_off_fin = -101.698 uV 10:45:08 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 59.601 uVs 10:45:08 Integrated Voltage Measurement... 10:45:17 Voltage Offset Correction: v_off_ini = -88.149 uV, v_off_fin = -75.950 uV 10:45:17 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -46.310 uVs 10:45:17 Integrated Voltage Measurement... 10:45:26 Voltage Offset Correction: v_off_ini = -89.117 uV, v_off_fin = -83.759 uV 10:45:26 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 46.135 uVs 10:45:28 I2X Average and RMS of Measurements VT = -54.048 +- 8.780 uVs I2X = 4.972 +- 0.808 uTm^2 10:45:31 Wire Position: X0 = 0.0000 mm, Y0 = 1.0000 mm 10:45:33 Integrated Voltage Measurement... 10:45:42 Voltage Offset Correction: v_off_ini = -82.189 uV, v_off_fin = -82.796 uV 10:45:42 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -234.431 uVs 10:45:42 Integrated Voltage Measurement... 10:45:51 Voltage Offset Correction: v_off_ini = -85.145 uV, v_off_fin = -81.403 uV 10:45:51 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 217.908 uVs 10:45:51 Integrated Voltage Measurement... 10:46:00 Voltage Offset Correction: v_off_ini = -104.825 uV, v_off_fin = -99.487 uV 10:46:00 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -226.326 uVs 10:46:00 Integrated Voltage Measurement... 10:46:09 Voltage Offset Correction: v_off_ini = -98.269 uV, v_off_fin = -99.408 uV 10:46:09 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 219.017 uVs 10:46:09 Integrated Voltage Measurement... 10:46:18 Voltage Offset Correction: v_off_ini = -105.584 uV, v_off_fin = -88.210 uV 10:46:18 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -217.706 uVs 10:46:18 Integrated Voltage Measurement... 10:46:27 Voltage Offset Correction: v_off_ini = -104.662 uV, v_off_fin = -81.613 uV 10:46:27 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 239.190 uVs 10:46:30 I2Y Average and RMS of Measurements VT = -225.763 +- 2.376 uVs I2Y = -20.770 +- 0.219 uTm^2 10:46:46 Wire Position: X0 = -2.0000 mm, Y0 = 0.0000 mm 10:46:50 Integrated Voltage Measurement... 10:46:59 Voltage Offset Correction: v_off_ini = -112.677 uV, v_off_fin = -113.860 uV 10:46:59 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -23.828 uVs 10:46:59 Integrated Voltage Measurement... 10:47:08 Voltage Offset Correction: v_off_ini = -99.374 uV, v_off_fin = -91.745 uV 10:47:08 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -5.806 uVs 10:47:08 Integrated Voltage Measurement... 10:47:17 Voltage Offset Correction: v_off_ini = -100.844 uV, v_off_fin = -111.411 uV 10:47:17 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -37.636 uVs 10:47:17 Integrated Voltage Measurement... 10:47:26 Voltage Offset Correction: v_off_ini = -108.767 uV, v_off_fin = -103.797 uV 10:47:26 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -2.393 uVs 10:47:26 Integrated Voltage Measurement... 10:47:35 Voltage Offset Correction: v_off_ini = -106.776 uV, v_off_fin = -102.000 uV 10:47:35 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -18.904 uVs 10:47:35 Integrated Voltage Measurement... 10:47:44 Voltage Offset Correction: v_off_ini = -106.612 uV, v_off_fin = -104.209 uV 10:47:44 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 7.311 uVs 10:47:48 I1X Average and RMS of Measurements VT = -13.247 +- 3.517 uVs I1X = 2.649 +- 0.703 uTm 10:47:51 Wire Position: X0 = -2.0000 mm, Y0 = 0.0000 mm 10:47:55 Integrated Voltage Measurement... 10:48:04 Voltage Offset Correction: v_off_ini = -128.365 uV, v_off_fin = -121.001 uV 10:48:04 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -160.336 uVs 10:48:04 Integrated Voltage Measurement... 10:48:13 Voltage Offset Correction: v_off_ini = -108.976 uV, v_off_fin = -102.838 uV 10:48:13 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 178.208 uVs 10:48:13 Integrated Voltage Measurement... 10:48:22 Voltage Offset Correction: v_off_ini = -95.322 uV, v_off_fin = -121.847 uV 10:48:22 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -159.625 uVs 10:48:22 Integrated Voltage Measurement... 10:48:31 Voltage Offset Correction: v_off_ini = -114.421 uV, v_off_fin = -104.005 uV 10:48:31 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 160.970 uVs 10:48:31 Integrated Voltage Measurement... 10:48:40 Voltage Offset Correction: v_off_ini = -105.035 uV, v_off_fin = -112.382 uV 10:48:40 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -176.477 uVs 10:48:40 Integrated Voltage Measurement... 10:48:49 Voltage Offset Correction: v_off_ini = -108.923 uV, v_off_fin = -94.333 uV 10:48:49 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 171.801 uVs 10:48:53 I1Y Average and RMS of Measurements VT = -167.903 +- 5.733 uVs I1Y = -33.581 +- 1.147 uTm 10:48:58 Wire Position: X0 = -2.0000 mm, Y0 = 0.0000 mm 10:49:01 Integrated Voltage Measurement... 10:49:10 Voltage Offset Correction: v_off_ini = -114.718 uV, v_off_fin = -107.981 uV 10:49:10 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -10.010 uVs 10:49:10 Integrated Voltage Measurement... 10:49:19 Voltage Offset Correction: v_off_ini = -122.340 uV, v_off_fin = -122.536 uV 10:49:19 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 17.587 uVs 10:49:19 Integrated Voltage Measurement... 10:49:28 Voltage Offset Correction: v_off_ini = -122.279 uV, v_off_fin = -127.862 uV 10:49:28 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -4.826 uVs 10:49:28 Integrated Voltage Measurement... 10:49:37 Voltage Offset Correction: v_off_ini = -118.740 uV, v_off_fin = -129.936 uV 10:49:37 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 16.726 uVs 10:49:37 Integrated Voltage Measurement... 10:49:46 Voltage Offset Correction: v_off_ini = -126.561 uV, v_off_fin = -118.009 uV 10:49:46 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -24.937 uVs 10:49:46 Integrated Voltage Measurement... 10:49:55 Voltage Offset Correction: v_off_ini = -129.785 uV, v_off_fin = -117.896 uV 10:49:55 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 11.377 uVs 10:49:58 I2X Average and RMS of Measurements VT = -14.244 +- 3.030 uVs I2X = 1.310 +- 0.279 uTm^2 10:50:03 Wire Position: X0 = -2.0000 mm, Y0 = 0.0000 mm 10:50:05 Integrated Voltage Measurement... 10:50:14 Voltage Offset Correction: v_off_ini = -135.714 uV, v_off_fin = -130.065 uV 10:50:14 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -180.192 uVs 10:50:14 Integrated Voltage Measurement... 10:50:23 Voltage Offset Correction: v_off_ini = -131.834 uV, v_off_fin = -132.479 uV 10:50:23 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 170.775 uVs 10:50:23 Integrated Voltage Measurement... 10:50:32 Voltage Offset Correction: v_off_ini = -139.003 uV, v_off_fin = -122.790 uV 10:50:32 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -182.342 uVs 10:50:32 Integrated Voltage Measurement... 10:50:41 Voltage Offset Correction: v_off_ini = -137.599 uV, v_off_fin = -121.897 uV 10:50:41 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 189.044 uVs 10:50:41 Integrated Voltage Measurement... 10:50:50 Voltage Offset Correction: v_off_ini = -124.707 uV, v_off_fin = -124.212 uV 10:50:50 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -187.429 uVs 10:50:50 Integrated Voltage Measurement... 10:50:59 Voltage Offset Correction: v_off_ini = -134.786 uV, v_off_fin = -150.699 uV 10:50:59 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 136.354 uVs 10:51:02 I2Y Average and RMS of Measurements VT = -174.356 +- 9.750 uVs I2Y = -16.041 +- 0.897 uTm^2 10:51:06 Wire Position: X0 = -1.0000 mm, Y0 = 0.0000 mm 10:51:11 Integrated Voltage Measurement... 10:51:20 Voltage Offset Correction: v_off_ini = -147.442 uV, v_off_fin = -142.122 uV 10:51:20 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -33.981 uVs 10:51:20 Integrated Voltage Measurement... 10:51:29 Voltage Offset Correction: v_off_ini = -152.248 uV, v_off_fin = -150.549 uV 10:51:29 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 34.525 uVs 10:51:29 Integrated Voltage Measurement... 10:51:38 Voltage Offset Correction: v_off_ini = -157.506 uV, v_off_fin = -162.823 uV 10:51:38 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -40.415 uVs 10:51:38 Integrated Voltage Measurement... 10:51:47 Voltage Offset Correction: v_off_ini = -159.073 uV, v_off_fin = -137.763 uV 10:51:47 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 11.348 uVs 10:51:47 Integrated Voltage Measurement... 10:51:56 Voltage Offset Correction: v_off_ini = -146.157 uV, v_off_fin = -161.866 uV 10:51:56 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -33.039 uVs 10:51:56 Integrated Voltage Measurement... 10:52:05 Voltage Offset Correction: v_off_ini = -142.677 uV, v_off_fin = -163.508 uV 10:52:05 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 18.016 uVs 10:52:09 I1X Average and RMS of Measurements VT = -28.554 +- 4.032 uVs I1X = 5.711 +- 0.806 uTm 10:52:12 Wire Position: X0 = -1.0000 mm, Y0 = 0.0000 mm 10:52:16 Integrated Voltage Measurement... 10:52:25 Voltage Offset Correction: v_off_ini = -135.434 uV, v_off_fin = -150.367 uV 10:52:25 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -153.901 uVs 10:52:25 Integrated Voltage Measurement... 10:52:34 Voltage Offset Correction: v_off_ini = -157.830 uV, v_off_fin = -139.701 uV 10:52:34 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 168.370 uVs 10:52:34 Integrated Voltage Measurement... 10:52:43 Voltage Offset Correction: v_off_ini = -156.384 uV, v_off_fin = -151.978 uV 10:52:43 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -179.385 uVs 10:52:43 Integrated Voltage Measurement... 10:52:52 Voltage Offset Correction: v_off_ini = -136.965 uV, v_off_fin = -131.737 uV 10:52:52 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 153.725 uVs 10:52:52 Integrated Voltage Measurement... 10:53:01 Voltage Offset Correction: v_off_ini = -142.013 uV, v_off_fin = -147.919 uV 10:53:01 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -161.481 uVs 10:53:01 Integrated Voltage Measurement... 10:53:10 Voltage Offset Correction: v_off_ini = -128.861 uV, v_off_fin = -121.202 uV 10:53:10 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 164.527 uVs 10:53:14 I1Y Average and RMS of Measurements VT = -163.565 +- 2.248 uVs I1Y = -32.713 +- 0.450 uTm 10:53:19 Wire Position: X0 = -1.0000 mm, Y0 = 0.0000 mm 10:53:22 Integrated Voltage Measurement... 10:53:31 Voltage Offset Correction: v_off_ini = -123.854 uV, v_off_fin = -138.168 uV 10:53:31 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -45.433 uVs 10:53:31 Integrated Voltage Measurement... 10:53:40 Voltage Offset Correction: v_off_ini = -131.285 uV, v_off_fin = -140.378 uV 10:53:40 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 50.954 uVs 10:53:40 Integrated Voltage Measurement... 10:53:49 Voltage Offset Correction: v_off_ini = -121.452 uV, v_off_fin = -134.395 uV 10:53:49 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -38.669 uVs 10:53:49 Integrated Voltage Measurement... 10:53:58 Voltage Offset Correction: v_off_ini = -129.261 uV, v_off_fin = -152.372 uV 10:53:58 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 40.735 uVs 10:53:58 Integrated Voltage Measurement... 10:54:07 Voltage Offset Correction: v_off_ini = -134.522 uV, v_off_fin = -132.384 uV 10:54:07 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -39.978 uVs 10:54:07 Integrated Voltage Measurement... 10:54:16 Voltage Offset Correction: v_off_ini = -142.139 uV, v_off_fin = -131.745 uV 10:54:16 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 26.998 uVs 10:54:18 I2X Average and RMS of Measurements VT = -40.461 +- 6.027 uVs I2X = 3.722 +- 0.555 uTm^2 10:54:24 Wire Position: X0 = -1.0000 mm, Y0 = 0.0000 mm 10:54:26 Integrated Voltage Measurement... 10:54:35 Voltage Offset Correction: v_off_ini = -138.197 uV, v_off_fin = -139.311 uV 10:54:35 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -184.189 uVs 10:54:35 Integrated Voltage Measurement... 10:54:44 Voltage Offset Correction: v_off_ini = -116.459 uV, v_off_fin = -129.232 uV 10:54:44 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 163.695 uVs 10:54:44 Integrated Voltage Measurement... 10:54:53 Voltage Offset Correction: v_off_ini = -121.537 uV, v_off_fin = -128.749 uV 10:54:53 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -170.409 uVs 10:54:53 Integrated Voltage Measurement... 10:55:02 Voltage Offset Correction: v_off_ini = -138.600 uV, v_off_fin = -134.727 uV 10:55:02 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 177.370 uVs 10:55:02 Integrated Voltage Measurement... 10:55:11 Voltage Offset Correction: v_off_ini = -140.311 uV, v_off_fin = -132.120 uV 10:55:11 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -170.494 uVs 10:55:11 Integrated Voltage Measurement... 10:55:20 Voltage Offset Correction: v_off_ini = -127.543 uV, v_off_fin = -135.893 uV 10:55:20 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 175.231 uVs 10:55:23 I2Y Average and RMS of Measurements VT = -173.565 +- 0.497 uVs I2Y = -15.968 +- 0.046 uTm^2 10:55:27 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:55:32 Integrated Voltage Measurement... 10:55:41 Voltage Offset Correction: v_off_ini = -140.937 uV, v_off_fin = -139.522 uV 10:55:41 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -34.377 uVs 10:55:41 Integrated Voltage Measurement... 10:55:50 Voltage Offset Correction: v_off_ini = -145.506 uV, v_off_fin = -141.519 uV 10:55:50 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 31.016 uVs 10:55:50 Integrated Voltage Measurement... 10:55:59 Voltage Offset Correction: v_off_ini = -132.901 uV, v_off_fin = -149.440 uV 10:55:59 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -42.706 uVs 10:55:59 Integrated Voltage Measurement... 10:56:08 Voltage Offset Correction: v_off_ini = -134.973 uV, v_off_fin = -127.717 uV 10:56:08 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 32.362 uVs 10:56:08 Integrated Voltage Measurement... 10:56:17 Voltage Offset Correction: v_off_ini = -142.320 uV, v_off_fin = -142.572 uV 10:56:17 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -30.656 uVs 10:56:17 Integrated Voltage Measurement... 10:56:26 Voltage Offset Correction: v_off_ini = -130.867 uV, v_off_fin = -124.457 uV 10:56:26 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 56.813 uVs 10:56:30 I1X Average and RMS of Measurements VT = -37.989 +- 4.518 uVs I1X = 7.598 +- 0.904 uTm 10:56:33 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:56:37 Integrated Voltage Measurement... 10:56:46 Voltage Offset Correction: v_off_ini = -154.035 uV, v_off_fin = -159.338 uV 10:56:46 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -164.147 uVs 10:56:46 Integrated Voltage Measurement... 10:56:55 Voltage Offset Correction: v_off_ini = -131.386 uV, v_off_fin = -119.258 uV 10:56:55 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.4999 mm dXY = -0.4999 mm VT = 169.856 uVs 10:56:55 Integrated Voltage Measurement... 10:57:04 Voltage Offset Correction: v_off_ini = -134.162 uV, v_off_fin = -137.440 uV 10:57:04 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.4999 mm dXY = 0.4999 mm VT = -178.104 uVs 10:57:04 Integrated Voltage Measurement... 10:57:13 Voltage Offset Correction: v_off_ini = -133.012 uV, v_off_fin = -133.326 uV 10:57:13 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.4999 mm dXY = -0.4999 mm VT = 163.843 uVs 10:57:13 Integrated Voltage Measurement... 10:57:22 Voltage Offset Correction: v_off_ini = -127.418 uV, v_off_fin = -141.605 uV 10:57:22 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -183.743 uVs 10:57:22 Integrated Voltage Measurement... 10:57:31 Voltage Offset Correction: v_off_ini = -150.628 uV, v_off_fin = -133.201 uV 10:57:31 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 164.663 uVs 10:57:35 I1Y Average and RMS of Measurements VT = -170.726 +- 2.945 uVs I1Y = -34.149 +- 0.588 uTm 10:57:40 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:57:43 Integrated Voltage Measurement... 10:57:52 Voltage Offset Correction: v_off_ini = -144.839 uV, v_off_fin = -135.671 uV 10:57:52 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -63.996 uVs 10:57:52 Integrated Voltage Measurement... 10:58:01 Voltage Offset Correction: v_off_ini = -156.700 uV, v_off_fin = -175.191 uV 10:58:01 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 47.451 uVs 10:58:01 Integrated Voltage Measurement... 10:58:10 Voltage Offset Correction: v_off_ini = -163.227 uV, v_off_fin = -149.490 uV 10:58:10 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -49.928 uVs 10:58:10 Integrated Voltage Measurement... 10:58:19 Voltage Offset Correction: v_off_ini = -159.429 uV, v_off_fin = -158.811 uV 10:58:19 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 45.379 uVs 10:58:19 Integrated Voltage Measurement... 10:58:28 Voltage Offset Correction: v_off_ini = -151.197 uV, v_off_fin = -171.660 uV 10:58:28 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -41.284 uVs 10:58:28 Integrated Voltage Measurement... 10:58:37 Voltage Offset Correction: v_off_ini = -164.860 uV, v_off_fin = -169.839 uV 10:58:37 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 16.670 uVs 10:58:39 I2X Average and RMS of Measurements VT = -44.118 +- 11.202 uVs I2X = 4.059 +- 1.031 uTm^2 10:58:45 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:58:47 Integrated Voltage Measurement... 10:58:56 Voltage Offset Correction: v_off_ini = -157.024 uV, v_off_fin = -165.263 uV 10:58:56 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -202.340 uVs 10:58:56 Integrated Voltage Measurement... 10:59:05 Voltage Offset Correction: v_off_ini = -179.922 uV, v_off_fin = -172.469 uV 10:59:05 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 178.269 uVs 10:59:05 Integrated Voltage Measurement... 10:59:14 Voltage Offset Correction: v_off_ini = -171.876 uV, v_off_fin = -183.966 uV 10:59:14 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -173.398 uVs 10:59:14 Integrated Voltage Measurement... 10:59:23 Voltage Offset Correction: v_off_ini = -167.684 uV, v_off_fin = -168.563 uV 10:59:23 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = 183.080 uVs 10:59:23 Integrated Voltage Measurement... 10:59:32 Voltage Offset Correction: v_off_ini = -171.599 uV, v_off_fin = -176.934 uV 10:59:32 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -199.589 uVs 10:59:32 Integrated Voltage Measurement... 10:59:41 Voltage Offset Correction: v_off_ini = -179.948 uV, v_off_fin = -178.846 uV 10:59:41 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 193.704 uVs 10:59:44 I2Y Average and RMS of Measurements VT = -188.397 +- 7.635 uVs I2Y = -17.333 +- 0.702 uTm^2 10:59:48 Wire Position: X0 = 1.0000 mm, Y0 = 0.0000 mm 10:59:53 Integrated Voltage Measurement... 11:00:02 Voltage Offset Correction: v_off_ini = -173.086 uV, v_off_fin = -171.841 uV 11:00:02 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -42.982 uVs 11:00:02 Integrated Voltage Measurement... 11:00:11 Voltage Offset Correction: v_off_ini = -160.349 uV, v_off_fin = -161.587 uV 11:00:11 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 56.753 uVs 11:00:11 Integrated Voltage Measurement... 11:00:20 Voltage Offset Correction: v_off_ini = -157.020 uV, v_off_fin = -157.260 uV 11:00:20 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -81.131 uVs 11:00:20 Integrated Voltage Measurement... 11:00:29 Voltage Offset Correction: v_off_ini = -170.434 uV, v_off_fin = -159.718 uV 11:00:29 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 51.980 uVs 11:00:29 Integrated Voltage Measurement... 11:00:38 Voltage Offset Correction: v_off_ini = -176.375 uV, v_off_fin = -170.302 uV 11:00:38 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -63.954 uVs 11:00:38 Integrated Voltage Measurement... 11:00:47 Voltage Offset Correction: v_off_ini = -158.676 uV, v_off_fin = -155.622 uV 11:00:47 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 51.925 uVs 11:00:51 I1X Average and RMS of Measurements VT = -58.121 +- 6.814 uVs I1X = 11.624 +- 1.363 uTm 11:00:54 Wire Position: X0 = 1.0000 mm, Y0 = 0.0000 mm 11:00:58 Integrated Voltage Measurement... 11:01:07 Voltage Offset Correction: v_off_ini = -176.761 uV, v_off_fin = -177.500 uV 11:01:07 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -160.217 uVs 11:01:07 Integrated Voltage Measurement... 11:01:16 Voltage Offset Correction: v_off_ini = -146.937 uV, v_off_fin = -147.640 uV 11:01:16 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 160.783 uVs 11:01:16 Integrated Voltage Measurement... 11:01:25 Voltage Offset Correction: v_off_ini = -154.920 uV, v_off_fin = -167.060 uV 11:01:25 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -162.753 uVs 11:01:25 Integrated Voltage Measurement... 11:01:34 Voltage Offset Correction: v_off_ini = -184.920 uV, v_off_fin = -171.388 uV 11:01:34 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 163.763 uVs 11:01:34 Integrated Voltage Measurement... 11:01:43 Voltage Offset Correction: v_off_ini = -156.646 uV, v_off_fin = -170.966 uV 11:01:43 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -190.242 uVs 11:01:43 Integrated Voltage Measurement... 11:01:52 Voltage Offset Correction: v_off_ini = -162.325 uV, v_off_fin = -151.408 uV 11:01:52 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 182.454 uVs 11:01:56 I1Y Average and RMS of Measurements VT = -170.035 +- 11.590 uVs I1Y = -34.007 +- 2.318 uTm 11:02:01 Wire Position: X0 = 1.0000 mm, Y0 = 0.0000 mm 11:02:04 Integrated Voltage Measurement... 11:02:13 Voltage Offset Correction: v_off_ini = -159.494 uV, v_off_fin = -144.817 uV 11:02:13 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -59.322 uVs 11:02:13 Integrated Voltage Measurement... 11:02:22 Voltage Offset Correction: v_off_ini = -140.709 uV, v_off_fin = -160.277 uV 11:02:22 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 52.163 uVs 11:02:22 Integrated Voltage Measurement... 11:02:31 Voltage Offset Correction: v_off_ini = -144.810 uV, v_off_fin = -142.587 uV 11:02:31 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -81.646 uVs 11:02:31 Integrated Voltage Measurement... 11:02:40 Voltage Offset Correction: v_off_ini = -180.002 uV, v_off_fin = -163.136 uV 11:02:40 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 62.947 uVs 11:02:40 Integrated Voltage Measurement... 11:02:49 Voltage Offset Correction: v_off_ini = -151.891 uV, v_off_fin = -152.061 uV 11:02:49 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -72.011 uVs 11:02:49 Integrated Voltage Measurement... 11:02:58 Voltage Offset Correction: v_off_ini = -157.275 uV, v_off_fin = -149.344 uV 11:02:58 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 74.379 uVs 11:03:00 I2X Average and RMS of Measurements VT = -67.078 +- 8.024 uVs I2X = 6.171 +- 0.738 uTm^2 11:03:05 Wire Position: X0 = 1.0000 mm, Y0 = 0.0000 mm 11:03:08 Integrated Voltage Measurement... 11:03:17 Voltage Offset Correction: v_off_ini = -160.963 uV, v_off_fin = -150.926 uV 11:03:17 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -199.507 uVs 11:03:17 Integrated Voltage Measurement... 11:03:26 Voltage Offset Correction: v_off_ini = -147.912 uV, v_off_fin = -139.192 uV 11:03:26 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = 183.560 uVs 11:03:26 Integrated Voltage Measurement... 11:03:35 Voltage Offset Correction: v_off_ini = -135.431 uV, v_off_fin = -142.063 uV 11:03:35 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -157.241 uVs 11:03:35 Integrated Voltage Measurement... 11:03:44 Voltage Offset Correction: v_off_ini = -135.616 uV, v_off_fin = -134.031 uV 11:03:44 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 180.166 uVs 11:03:44 Integrated Voltage Measurement... 11:03:53 Voltage Offset Correction: v_off_ini = -140.945 uV, v_off_fin = -154.928 uV 11:03:53 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -202.458 uVs 11:03:53 Integrated Voltage Measurement... 11:04:02 Voltage Offset Correction: v_off_ini = -153.389 uV, v_off_fin = -145.951 uV 11:04:02 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = 185.438 uVs 11:04:05 I2Y Average and RMS of Measurements VT = -184.728 +- 11.374 uVs I2Y = -16.996 +- 1.047 uTm^2 11:04:09 Wire Position: X0 = 2.0000 mm, Y0 = 0.0000 mm 11:04:14 Integrated Voltage Measurement... 11:04:23 Voltage Offset Correction: v_off_ini = -157.541 uV, v_off_fin = -161.812 uV 11:04:23 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -72.145 uVs 11:04:23 Integrated Voltage Measurement... 11:04:32 Voltage Offset Correction: v_off_ini = -148.100 uV, v_off_fin = -159.907 uV 11:04:32 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 85.069 uVs 11:04:32 Integrated Voltage Measurement... 11:04:41 Voltage Offset Correction: v_off_ini = -150.896 uV, v_off_fin = -154.054 uV 11:04:41 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -78.356 uVs 11:04:41 Integrated Voltage Measurement... 11:04:50 Voltage Offset Correction: v_off_ini = -144.843 uV, v_off_fin = -150.318 uV 11:04:50 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 59.412 uVs 11:04:50 Integrated Voltage Measurement... 11:04:59 Voltage Offset Correction: v_off_ini = -143.961 uV, v_off_fin = -146.276 uV 11:04:59 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -57.797 uVs 11:04:59 Integrated Voltage Measurement... 11:05:08 Voltage Offset Correction: v_off_ini = -139.851 uV, v_off_fin = -151.898 uV 11:05:08 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 72.051 uVs 11:05:12 I1X Average and RMS of Measurements VT = -70.805 +- 5.749 uVs I1X = 14.161 +- 1.150 uTm 11:05:14 Wire Position: X0 = 2.0000 mm, Y0 = 0.0000 mm 11:05:19 Integrated Voltage Measurement... 11:05:28 Voltage Offset Correction: v_off_ini = -140.962 uV, v_off_fin = -143.407 uV 11:05:28 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.4999 mm dXY = 0.4999 mm VT = -177.226 uVs 11:05:28 Integrated Voltage Measurement... 11:05:37 Voltage Offset Correction: v_off_ini = -130.063 uV, v_off_fin = -132.445 uV 11:05:37 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 159.698 uVs 11:05:37 Integrated Voltage Measurement... 11:05:46 Voltage Offset Correction: v_off_ini = -132.790 uV, v_off_fin = -138.555 uV 11:05:46 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -181.119 uVs 11:05:46 Integrated Voltage Measurement... 11:05:55 Voltage Offset Correction: v_off_ini = -137.119 uV, v_off_fin = -130.808 uV 11:05:55 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 170.896 uVs 11:05:55 Integrated Voltage Measurement... 11:06:04 Voltage Offset Correction: v_off_ini = -115.095 uV, v_off_fin = -133.372 uV 11:06:04 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -157.074 uVs 11:06:04 Integrated Voltage Measurement... 11:06:13 Voltage Offset Correction: v_off_ini = -132.531 uV, v_off_fin = -123.169 uV 11:06:13 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 174.477 uVs 11:06:17 I1Y Average and RMS of Measurements VT = -170.082 +- 4.331 uVs I1Y = -34.018 +- 0.866 uTm 11:06:22 Wire Position: X0 = 2.0000 mm, Y0 = 0.0000 mm 11:06:25 Integrated Voltage Measurement... 11:06:34 Voltage Offset Correction: v_off_ini = -131.590 uV, v_off_fin = -123.313 uV 11:06:34 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -90.633 uVs 11:06:34 Integrated Voltage Measurement... 11:06:43 Voltage Offset Correction: v_off_ini = -121.949 uV, v_off_fin = -131.567 uV 11:06:43 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 79.600 uVs 11:06:43 Integrated Voltage Measurement... 11:06:52 Voltage Offset Correction: v_off_ini = -137.928 uV, v_off_fin = -139.284 uV 11:06:52 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -92.540 uVs 11:06:52 Integrated Voltage Measurement... 11:07:01 Voltage Offset Correction: v_off_ini = -139.658 uV, v_off_fin = -147.538 uV 11:07:01 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 74.645 uVs 11:07:01 Integrated Voltage Measurement... 11:07:10 Voltage Offset Correction: v_off_ini = -154.931 uV, v_off_fin = -146.722 uV 11:07:10 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -58.892 uVs 11:07:10 Integrated Voltage Measurement... 11:07:19 Voltage Offset Correction: v_off_ini = -148.756 uV, v_off_fin = -148.451 uV 11:07:19 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 71.814 uVs 11:07:21 I2X Average and RMS of Measurements VT = -78.020 +- 8.979 uVs I2X = 7.178 +- 0.826 uTm^2 11:07:26 Wire Position: X0 = 2.0000 mm, Y0 = 0.0000 mm 11:07:29 Integrated Voltage Measurement... 11:07:38 Voltage Offset Correction: v_off_ini = -143.164 uV, v_off_fin = -140.150 uV 11:07:38 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -186.350 uVs 11:07:38 Integrated Voltage Measurement... 11:07:47 Voltage Offset Correction: v_off_ini = -144.590 uV, v_off_fin = -139.133 uV 11:07:47 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 189.371 uVs 11:07:47 Integrated Voltage Measurement... 11:07:56 Voltage Offset Correction: v_off_ini = -154.044 uV, v_off_fin = -158.304 uV 11:07:56 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -197.197 uVs 11:07:56 Integrated Voltage Measurement... 11:08:05 Voltage Offset Correction: v_off_ini = -161.629 uV, v_off_fin = -160.619 uV 11:08:05 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 198.207 uVs 11:08:05 Integrated Voltage Measurement... 11:08:14 Voltage Offset Correction: v_off_ini = -151.381 uV, v_off_fin = -148.314 uV 11:08:14 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -193.900 uVs 11:08:14 Integrated Voltage Measurement... 11:08:23 Voltage Offset Correction: v_off_ini = -132.510 uV, v_off_fin = -130.487 uV 11:08:23 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 185.493 uVs 11:08:26 I2Y Average and RMS of Measurements VT = -191.753 +- 4.273 uVs I2Y = -17.641 +- 0.393 uTm^2 11:08:33 Wire Position: X0 = 0.0000 mm, Y0 = -1.0000 mm 11:08:38 Integrated Voltage Measurement... 11:08:47 Voltage Offset Correction: v_off_ini = -131.394 uV, v_off_fin = -120.099 uV 11:08:47 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -45.489 uVs 11:08:47 Integrated Voltage Measurement... 11:08:56 Voltage Offset Correction: v_off_ini = -123.310 uV, v_off_fin = -129.111 uV 11:08:56 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 34.488 uVs 11:08:56 Integrated Voltage Measurement... 11:09:05 Voltage Offset Correction: v_off_ini = -133.619 uV, v_off_fin = -144.077 uV 11:09:05 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -41.612 uVs 11:09:05 Integrated Voltage Measurement... 11:09:14 Voltage Offset Correction: v_off_ini = -141.095 uV, v_off_fin = -143.960 uV 11:09:14 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 50.475 uVs 11:09:14 Integrated Voltage Measurement... 11:09:23 Voltage Offset Correction: v_off_ini = -128.560 uV, v_off_fin = -132.738 uV 11:09:23 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -50.694 uVs 11:09:23 Integrated Voltage Measurement... 11:09:32 Voltage Offset Correction: v_off_ini = -136.536 uV, v_off_fin = -141.635 uV 11:09:32 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 45.195 uVs 11:09:36 I1X Average and RMS of Measurements VT = -44.659 +- 3.392 uVs I1X = 8.932 +- 0.678 uTm 11:09:38 Wire Position: X0 = 0.0000 mm, Y0 = -1.0000 mm 11:09:43 Integrated Voltage Measurement... 11:09:52 Voltage Offset Correction: v_off_ini = -136.292 uV, v_off_fin = -138.681 uV 11:09:52 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -167.314 uVs 11:09:52 Integrated Voltage Measurement... 11:10:01 Voltage Offset Correction: v_off_ini = -139.610 uV, v_off_fin = -136.307 uV 11:10:01 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 160.443 uVs 11:10:01 Integrated Voltage Measurement... 11:10:10 Voltage Offset Correction: v_off_ini = -138.645 uV, v_off_fin = -142.829 uV 11:10:10 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -168.364 uVs 11:10:10 Integrated Voltage Measurement... 11:10:19 Voltage Offset Correction: v_off_ini = -144.272 uV, v_off_fin = -135.013 uV 11:10:19 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 165.144 uVs 11:10:19 Integrated Voltage Measurement... 11:10:28 Voltage Offset Correction: v_off_ini = -155.984 uV, v_off_fin = -160.977 uV 11:10:28 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -155.525 uVs 11:10:28 Integrated Voltage Measurement... 11:10:37 Voltage Offset Correction: v_off_ini = -147.508 uV, v_off_fin = -141.000 uV 11:10:37 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 177.167 uVs 11:10:41 I1Y Average and RMS of Measurements VT = -165.660 +- 1.270 uVs I1Y = -33.132 +- 0.254 uTm 11:10:43 Wire Position: X0 = 0.0000 mm, Y0 = -1.0000 mm 11:10:46 Integrated Voltage Measurement... 11:10:55 Voltage Offset Correction: v_off_ini = -129.353 uV, v_off_fin = -150.061 uV 11:10:55 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -41.724 uVs 11:10:55 Integrated Voltage Measurement... 11:11:04 Voltage Offset Correction: v_off_ini = -151.289 uV, v_off_fin = -146.206 uV 11:11:04 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 49.401 uVs 11:11:04 Integrated Voltage Measurement... 11:11:13 Voltage Offset Correction: v_off_ini = -138.799 uV, v_off_fin = -132.235 uV 11:11:13 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -50.253 uVs 11:11:13 Integrated Voltage Measurement... 11:11:22 Voltage Offset Correction: v_off_ini = -139.550 uV, v_off_fin = -145.351 uV 11:11:22 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 41.160 uVs 11:11:22 Integrated Voltage Measurement... 11:11:31 Voltage Offset Correction: v_off_ini = -141.441 uV, v_off_fin = -149.000 uV 11:11:31 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -38.730 uVs 11:11:31 Integrated Voltage Measurement... 11:11:40 Voltage Offset Correction: v_off_ini = -154.199 uV, v_off_fin = -161.551 uV 11:11:40 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 50.949 uVs 11:11:43 I2X Average and RMS of Measurements VT = -45.369 +- 0.379 uVs I2X = 4.174 +- 0.035 uTm^2 11:11:45 Wire Position: X0 = 0.0000 mm, Y0 = -1.0000 mm 11:11:48 Integrated Voltage Measurement... 11:11:57 Voltage Offset Correction: v_off_ini = -144.648 uV, v_off_fin = -147.539 uV 11:11:57 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -156.819 uVs 11:11:57 Integrated Voltage Measurement... 11:12:06 Voltage Offset Correction: v_off_ini = -153.293 uV, v_off_fin = -158.589 uV 11:12:06 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 159.393 uVs 11:12:06 Integrated Voltage Measurement... 11:12:15 Voltage Offset Correction: v_off_ini = -164.397 uV, v_off_fin = -160.799 uV 11:12:15 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -177.347 uVs 11:12:15 Integrated Voltage Measurement... 11:12:24 Voltage Offset Correction: v_off_ini = -155.394 uV, v_off_fin = -129.575 uV 11:12:24 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 182.152 uVs 11:12:24 Integrated Voltage Measurement... 11:12:33 Voltage Offset Correction: v_off_ini = -146.976 uV, v_off_fin = -152.723 uV 11:12:33 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -151.836 uVs 11:12:33 Integrated Voltage Measurement... 11:12:42 Voltage Offset Correction: v_off_ini = -137.030 uV, v_off_fin = -141.544 uV 11:12:42 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 157.358 uVs 11:12:44 I2Y Average and RMS of Measurements VT = -164.151 +- 11.123 uVs I2Y = -15.102 +- 1.023 uTm^2 11:12:48 Wire Position: X0 = 0.0000 mm, Y0 = -0.5000 mm 11:12:53 Integrated Voltage Measurement... 11:13:02 Voltage Offset Correction: v_off_ini = -159.455 uV, v_off_fin = -154.128 uV 11:13:02 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -43.216 uVs 11:13:02 Integrated Voltage Measurement... 11:13:11 Voltage Offset Correction: v_off_ini = -157.172 uV, v_off_fin = -151.864 uV 11:13:11 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 29.357 uVs 11:13:11 Integrated Voltage Measurement... 11:13:20 Voltage Offset Correction: v_off_ini = -147.130 uV, v_off_fin = -142.066 uV 11:13:20 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -28.349 uVs 11:13:20 Integrated Voltage Measurement... 11:13:29 Voltage Offset Correction: v_off_ini = -145.754 uV, v_off_fin = -137.019 uV 11:13:29 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 39.487 uVs 11:13:29 Integrated Voltage Measurement... 11:13:38 Voltage Offset Correction: v_off_ini = -126.442 uV, v_off_fin = -154.166 uV 11:13:38 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -45.133 uVs 11:13:38 Integrated Voltage Measurement... 11:13:47 Voltage Offset Correction: v_off_ini = -155.975 uV, v_off_fin = -149.629 uV 11:13:47 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 44.229 uVs 11:13:51 I1X Average and RMS of Measurements VT = -38.295 +- 4.618 uVs I1X = 7.659 +- 0.924 uTm 11:13:53 Wire Position: X0 = 0.0000 mm, Y0 = -0.5000 mm 11:13:58 Integrated Voltage Measurement... 11:14:07 Voltage Offset Correction: v_off_ini = -159.669 uV, v_off_fin = -165.668 uV 11:14:07 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -173.724 uVs 11:14:07 Integrated Voltage Measurement... 11:14:16 Voltage Offset Correction: v_off_ini = -147.591 uV, v_off_fin = -146.945 uV 11:14:16 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 150.091 uVs 11:14:16 Integrated Voltage Measurement... 11:14:25 Voltage Offset Correction: v_off_ini = -157.951 uV, v_off_fin = -162.356 uV 11:14:25 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -171.456 uVs 11:14:25 Integrated Voltage Measurement... 11:14:34 Voltage Offset Correction: v_off_ini = -147.997 uV, v_off_fin = -149.893 uV 11:14:34 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 170.394 uVs 11:14:34 Integrated Voltage Measurement... 11:14:43 Voltage Offset Correction: v_off_ini = -160.808 uV, v_off_fin = -159.070 uV 11:14:43 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -161.799 uVs 11:14:43 Integrated Voltage Measurement... 11:14:52 Voltage Offset Correction: v_off_ini = -153.931 uV, v_off_fin = -150.147 uV 11:14:52 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 157.915 uVs 11:14:56 I1Y Average and RMS of Measurements VT = -164.230 +- 4.807 uVs I1Y = -32.846 +- 0.961 uTm 11:14:58 Wire Position: X0 = 0.0000 mm, Y0 = -0.5000 mm 11:15:01 Integrated Voltage Measurement... 11:15:10 Voltage Offset Correction: v_off_ini = -155.587 uV, v_off_fin = -160.592 uV 11:15:10 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -63.573 uVs 11:15:10 Integrated Voltage Measurement... 11:15:19 Voltage Offset Correction: v_off_ini = -161.032 uV, v_off_fin = -162.541 uV 11:15:19 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 61.069 uVs 11:15:19 Integrated Voltage Measurement... 11:15:28 Voltage Offset Correction: v_off_ini = -158.872 uV, v_off_fin = -162.792 uV 11:15:28 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -47.602 uVs 11:15:28 Integrated Voltage Measurement... 11:15:37 Voltage Offset Correction: v_off_ini = -172.226 uV, v_off_fin = -191.668 uV 11:15:37 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 52.347 uVs 11:15:37 Integrated Voltage Measurement... 11:15:46 Voltage Offset Correction: v_off_ini = -159.534 uV, v_off_fin = -155.213 uV 11:15:46 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -45.062 uVs 11:15:46 Integrated Voltage Measurement... 11:15:55 Voltage Offset Correction: v_off_ini = -159.518 uV, v_off_fin = -168.445 uV 11:15:55 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 39.834 uVs 11:15:58 I2X Average and RMS of Measurements VT = -51.581 +- 8.193 uVs I2X = 4.745 +- 0.754 uTm^2 11:16:00 Wire Position: X0 = 0.0000 mm, Y0 = -0.5000 mm 11:16:03 Integrated Voltage Measurement... 11:16:12 Voltage Offset Correction: v_off_ini = -165.997 uV, v_off_fin = -181.624 uV 11:16:12 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -182.587 uVs 11:16:12 Integrated Voltage Measurement... 11:16:21 Voltage Offset Correction: v_off_ini = -160.506 uV, v_off_fin = -152.782 uV 11:16:21 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 173.723 uVs 11:16:21 Integrated Voltage Measurement... 11:16:30 Voltage Offset Correction: v_off_ini = -170.257 uV, v_off_fin = -168.801 uV 11:16:30 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -202.270 uVs 11:16:30 Integrated Voltage Measurement... 11:16:39 Voltage Offset Correction: v_off_ini = -158.721 uV, v_off_fin = -164.271 uV 11:16:39 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 170.722 uVs 11:16:39 Integrated Voltage Measurement... 11:16:48 Voltage Offset Correction: v_off_ini = -164.835 uV, v_off_fin = -153.616 uV 11:16:48 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -162.345 uVs 11:16:48 Integrated Voltage Measurement... 11:16:57 Voltage Offset Correction: v_off_ini = -162.980 uV, v_off_fin = -163.316 uV 11:16:57 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 174.181 uVs 11:16:59 I2Y Average and RMS of Measurements VT = -177.638 +- 7.453 uVs I2Y = -16.343 +- 0.686 uTm^2 11:17:03 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:17:08 Integrated Voltage Measurement... 11:17:17 Voltage Offset Correction: v_off_ini = -159.569 uV, v_off_fin = -162.440 uV 11:17:17 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -74.284 uVs 11:17:17 Integrated Voltage Measurement... 11:17:26 Voltage Offset Correction: v_off_ini = -170.561 uV, v_off_fin = -152.711 uV 11:17:26 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 26.622 uVs 11:17:26 Integrated Voltage Measurement... 11:17:35 Voltage Offset Correction: v_off_ini = -159.406 uV, v_off_fin = -152.022 uV 11:17:35 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -56.554 uVs 11:17:35 Integrated Voltage Measurement... 11:17:44 Voltage Offset Correction: v_off_ini = -158.786 uV, v_off_fin = -151.264 uV 11:17:44 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 58.573 uVs 11:17:44 Integrated Voltage Measurement... 11:17:53 Voltage Offset Correction: v_off_ini = -159.726 uV, v_off_fin = -153.909 uV 11:17:53 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -38.253 uVs 11:17:53 Integrated Voltage Measurement... 11:18:02 Voltage Offset Correction: v_off_ini = -148.856 uV, v_off_fin = -144.489 uV 11:18:02 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 53.348 uVs 11:18:06 I1X Average and RMS of Measurements VT = -51.272 +- 4.837 uVs I1X = 10.254 +- 0.967 uTm 11:18:09 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:18:13 Integrated Voltage Measurement... 11:18:22 Voltage Offset Correction: v_off_ini = -149.534 uV, v_off_fin = -159.468 uV 11:18:22 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -177.571 uVs 11:18:22 Integrated Voltage Measurement... 11:18:31 Voltage Offset Correction: v_off_ini = -142.412 uV, v_off_fin = -139.732 uV 11:18:31 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 177.105 uVs 11:18:31 Integrated Voltage Measurement... 11:18:40 Voltage Offset Correction: v_off_ini = -144.781 uV, v_off_fin = -152.363 uV 11:18:40 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -174.053 uVs 11:18:40 Integrated Voltage Measurement... 11:18:49 Voltage Offset Correction: v_off_ini = -127.204 uV, v_off_fin = -125.256 uV 11:18:49 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 170.574 uVs 11:18:49 Integrated Voltage Measurement... 11:18:58 Voltage Offset Correction: v_off_ini = -138.140 uV, v_off_fin = -154.241 uV 11:18:58 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -175.601 uVs 11:18:58 Integrated Voltage Measurement... 11:19:07 Voltage Offset Correction: v_off_ini = -140.428 uV, v_off_fin = -132.340 uV 11:19:07 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 184.332 uVs 11:19:11 I1Y Average and RMS of Measurements VT = -176.539 +- 3.175 uVs I1Y = -35.308 +- 0.635 uTm 11:19:14 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:19:16 Integrated Voltage Measurement... 11:19:25 Voltage Offset Correction: v_off_ini = -138.300 uV, v_off_fin = -132.266 uV 11:19:25 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -51.653 uVs 11:19:25 Integrated Voltage Measurement... 11:19:34 Voltage Offset Correction: v_off_ini = -137.806 uV, v_off_fin = -137.491 uV 11:19:34 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 40.866 uVs 11:19:34 Integrated Voltage Measurement... 11:19:43 Voltage Offset Correction: v_off_ini = -144.869 uV, v_off_fin = -129.806 uV 11:19:43 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -54.925 uVs 11:19:43 Integrated Voltage Measurement... 11:19:52 Voltage Offset Correction: v_off_ini = -136.263 uV, v_off_fin = -134.649 uV 11:19:52 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 58.072 uVs 11:19:52 Integrated Voltage Measurement... 11:20:01 Voltage Offset Correction: v_off_ini = -124.393 uV, v_off_fin = -128.562 uV 11:20:01 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -53.114 uVs 11:20:01 Integrated Voltage Measurement... 11:20:10 Voltage Offset Correction: v_off_ini = -129.323 uV, v_off_fin = -137.776 uV 11:20:10 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 32.387 uVs 11:20:13 I2X Average and RMS of Measurements VT = -48.503 +- 5.833 uVs I2X = 4.462 +- 0.537 uTm^2 11:20:15 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:20:18 Integrated Voltage Measurement... 11:20:27 Voltage Offset Correction: v_off_ini = -139.617 uV, v_off_fin = -153.763 uV 11:20:27 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -194.593 uVs 11:20:27 Integrated Voltage Measurement... 11:20:36 Voltage Offset Correction: v_off_ini = -140.001 uV, v_off_fin = -142.080 uV 11:20:36 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 162.652 uVs 11:20:36 Integrated Voltage Measurement... 11:20:45 Voltage Offset Correction: v_off_ini = -146.161 uV, v_off_fin = -144.735 uV 11:20:45 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -184.243 uVs 11:20:45 Integrated Voltage Measurement... 11:20:54 Voltage Offset Correction: v_off_ini = -148.280 uV, v_off_fin = -134.503 uV 11:20:54 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 174.302 uVs 11:20:54 Integrated Voltage Measurement... 11:21:03 Voltage Offset Correction: v_off_ini = -133.889 uV, v_off_fin = -142.928 uV 11:21:03 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -183.172 uVs 11:21:03 Integrated Voltage Measurement... 11:21:12 Voltage Offset Correction: v_off_ini = -136.120 uV, v_off_fin = -140.178 uV 11:21:12 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 175.345 uVs 11:21:15 I2Y Average and RMS of Measurements VT = -179.051 +- 0.303 uVs I2Y = -16.473 +- 0.028 uTm^2 11:21:19 Wire Position: X0 = 0.0000 mm, Y0 = 0.5000 mm 11:21:23 Integrated Voltage Measurement... 11:21:32 Voltage Offset Correction: v_off_ini = -124.710 uV, v_off_fin = -138.126 uV 11:21:32 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.4999 mm dXY = -0.4999 mm VT = -34.346 uVs 11:21:32 Integrated Voltage Measurement... 11:21:41 Voltage Offset Correction: v_off_ini = -135.839 uV, v_off_fin = -132.071 uV 11:21:41 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 20.075 uVs 11:21:41 Integrated Voltage Measurement... 11:21:50 Voltage Offset Correction: v_off_ini = -135.309 uV, v_off_fin = -124.080 uV 11:21:50 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -36.644 uVs 11:21:50 Integrated Voltage Measurement... 11:21:59 Voltage Offset Correction: v_off_ini = -124.467 uV, v_off_fin = -117.834 uV 11:21:59 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = 36.111 uVs 11:21:59 Integrated Voltage Measurement... 11:22:08 Voltage Offset Correction: v_off_ini = -128.928 uV, v_off_fin = -123.894 uV 11:22:08 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -42.385 uVs 11:22:08 Integrated Voltage Measurement... 11:22:17 Voltage Offset Correction: v_off_ini = -135.504 uV, v_off_fin = -129.711 uV 11:22:17 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 41.516 uVs 11:22:21 I1X Average and RMS of Measurements VT = -35.180 +- 6.077 uVs I1X = 7.036 +- 1.215 uTm 11:22:24 Wire Position: X0 = 0.0000 mm, Y0 = 0.5000 mm 11:22:28 Integrated Voltage Measurement... 11:22:37 Voltage Offset Correction: v_off_ini = -122.562 uV, v_off_fin = -151.504 uV 11:22:37 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = -182.425 uVs 11:22:37 Integrated Voltage Measurement... 11:22:46 Voltage Offset Correction: v_off_ini = -141.305 uV, v_off_fin = -133.580 uV 11:22:46 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 178.566 uVs 11:22:46 Integrated Voltage Measurement... 11:22:55 Voltage Offset Correction: v_off_ini = -126.420 uV, v_off_fin = -130.163 uV 11:22:55 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -192.218 uVs 11:22:55 Integrated Voltage Measurement... 11:23:04 Voltage Offset Correction: v_off_ini = -121.096 uV, v_off_fin = -109.239 uV 11:23:04 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 178.101 uVs 11:23:04 Integrated Voltage Measurement... 11:23:13 Voltage Offset Correction: v_off_ini = -129.379 uV, v_off_fin = -152.669 uV 11:23:13 Integrated Voltage Measurement dXYus = 0.5001 mm, dXYds = 0.5001 mm dXY = 0.5001 mm VT = -176.111 uVs 11:23:13 Integrated Voltage Measurement... 11:23:22 Voltage Offset Correction: v_off_ini = -137.657 uV, v_off_fin = -124.709 uV 11:23:22 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 177.136 uVs 11:23:26 I1Y Average and RMS of Measurements VT = -180.759 +- 3.490 uVs I1Y = -36.151 +- 0.700 uTm 11:23:29 Wire Position: X0 = 0.0000 mm, Y0 = 0.5000 mm 11:23:31 Integrated Voltage Measurement... 11:23:40 Voltage Offset Correction: v_off_ini = -124.238 uV, v_off_fin = -136.559 uV 11:23:40 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -74.652 uVs 11:23:40 Integrated Voltage Measurement... 11:23:49 Voltage Offset Correction: v_off_ini = -136.871 uV, v_off_fin = -144.665 uV 11:23:49 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 38.419 uVs 11:23:49 Integrated Voltage Measurement... 11:23:58 Voltage Offset Correction: v_off_ini = -137.212 uV, v_off_fin = -131.648 uV 11:23:58 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -31.486 uVs 11:23:58 Integrated Voltage Measurement... 11:24:07 Voltage Offset Correction: v_off_ini = -139.497 uV, v_off_fin = -142.268 uV 11:24:07 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 32.521 uVs 11:24:07 Integrated Voltage Measurement... 11:24:16 Voltage Offset Correction: v_off_ini = -132.826 uV, v_off_fin = -119.559 uV 11:24:16 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -36.532 uVs 11:24:17 Integrated Voltage Measurement... 11:24:25 Voltage Offset Correction: v_off_ini = -127.357 uV, v_off_fin = -146.849 uV 11:24:26 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 21.683 uVs 11:24:28 I2X Average and RMS of Measurements VT = -39.215 +- 12.304 uVs I2X = 3.608 +- 1.132 uTm^2 11:24:30 Wire Position: X0 = 0.0000 mm, Y0 = 0.5000 mm 11:24:33 Integrated Voltage Measurement... 11:24:42 Voltage Offset Correction: v_off_ini = -129.299 uV, v_off_fin = -130.610 uV 11:24:42 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -186.495 uVs 11:24:42 Integrated Voltage Measurement... 11:24:51 Voltage Offset Correction: v_off_ini = -118.733 uV, v_off_fin = -117.874 uV 11:24:51 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 196.828 uVs 11:24:51 Integrated Voltage Measurement... 11:25:00 Voltage Offset Correction: v_off_ini = -143.025 uV, v_off_fin = -129.381 uV 11:25:00 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -163.946 uVs 11:25:00 Integrated Voltage Measurement... 11:25:09 Voltage Offset Correction: v_off_ini = -132.269 uV, v_off_fin = -131.175 uV 11:25:09 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 196.676 uVs 11:25:09 Integrated Voltage Measurement... 11:25:18 Voltage Offset Correction: v_off_ini = -136.602 uV, v_off_fin = -141.310 uV 11:25:18 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -188.334 uVs 11:25:18 Integrated Voltage Measurement... 11:25:27 Voltage Offset Correction: v_off_ini = -143.429 uV, v_off_fin = -139.353 uV 11:25:27 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 190.552 uVs 11:25:30 I2Y Average and RMS of Measurements VT = -187.138 +- 4.912 uVs I2Y = -17.217 +- 0.452 uTm^2 11:25:34 Wire Position: X0 = 0.0000 mm, Y0 = 1.0000 mm 11:25:38 Integrated Voltage Measurement... 11:25:47 Voltage Offset Correction: v_off_ini = -152.496 uV, v_off_fin = -136.988 uV 11:25:47 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -56.235 uVs 11:25:47 Integrated Voltage Measurement... 11:25:56 Voltage Offset Correction: v_off_ini = -123.346 uV, v_off_fin = -128.678 uV 11:25:56 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 27.491 uVs 11:25:56 Integrated Voltage Measurement... 11:26:05 Voltage Offset Correction: v_off_ini = -130.956 uV, v_off_fin = -141.750 uV 11:26:05 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -34.096 uVs 11:26:05 Integrated Voltage Measurement... 11:26:14 Voltage Offset Correction: v_off_ini = -133.301 uV, v_off_fin = -128.717 uV 11:26:14 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 45.598 uVs 11:26:14 Integrated Voltage Measurement... 11:26:23 Voltage Offset Correction: v_off_ini = -131.044 uV, v_off_fin = -150.315 uV 11:26:23 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -30.218 uVs 11:26:23 Integrated Voltage Measurement... 11:26:32 Voltage Offset Correction: v_off_ini = -146.543 uV, v_off_fin = -142.794 uV 11:26:32 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 65.194 uVs 11:26:36 I1X Average and RMS of Measurements VT = -43.139 +- 3.333 uVs I1X = 8.628 +- 0.667 uTm 11:26:39 Wire Position: X0 = 0.0000 mm, Y0 = 1.0000 mm 11:26:43 Integrated Voltage Measurement... 11:26:52 Voltage Offset Correction: v_off_ini = -149.256 uV, v_off_fin = -160.476 uV 11:26:52 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -175.925 uVs 11:26:52 Integrated Voltage Measurement... 11:27:01 Voltage Offset Correction: v_off_ini = -139.515 uV, v_off_fin = -131.234 uV 11:27:01 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 173.729 uVs 11:27:01 Integrated Voltage Measurement... 11:27:10 Voltage Offset Correction: v_off_ini = -139.931 uV, v_off_fin = -146.220 uV 11:27:10 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -195.576 uVs 11:27:10 Integrated Voltage Measurement... 11:27:19 Voltage Offset Correction: v_off_ini = -133.117 uV, v_off_fin = -123.572 uV 11:27:19 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 182.503 uVs 11:27:19 Integrated Voltage Measurement... 11:27:28 Voltage Offset Correction: v_off_ini = -120.755 uV, v_off_fin = -133.718 uV 11:27:28 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -183.872 uVs 11:27:28 Integrated Voltage Measurement... 11:27:37 Voltage Offset Correction: v_off_ini = -128.211 uV, v_off_fin = -123.732 uV 11:27:37 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 191.741 uVs 11:27:41 I1Y Average and RMS of Measurements VT = -183.891 +- 6.429 uVs I1Y = -36.778 +- 1.287 uTm 11:27:44 Wire Position: X0 = 0.0000 mm, Y0 = 1.0000 mm 11:27:46 Integrated Voltage Measurement... 11:27:55 Voltage Offset Correction: v_off_ini = -131.786 uV, v_off_fin = -120.007 uV 11:27:55 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -42.110 uVs 11:27:55 Integrated Voltage Measurement... 11:28:04 Voltage Offset Correction: v_off_ini = -105.410 uV, v_off_fin = -125.792 uV 11:28:04 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 18.321 uVs 11:28:04 Integrated Voltage Measurement... 11:28:13 Voltage Offset Correction: v_off_ini = -125.482 uV, v_off_fin = -127.617 uV 11:28:13 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -34.155 uVs 11:28:13 Integrated Voltage Measurement... 11:28:22 Voltage Offset Correction: v_off_ini = -137.260 uV, v_off_fin = -131.783 uV 11:28:22 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 34.396 uVs 11:28:23 Integrated Voltage Measurement... 11:28:31 Voltage Offset Correction: v_off_ini = -118.210 uV, v_off_fin = -121.221 uV 11:28:32 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -60.755 uVs 11:28:32 Integrated Voltage Measurement... 11:28:41 Voltage Offset Correction: v_off_ini = -126.514 uV, v_off_fin = -120.441 uV 11:28:41 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 36.638 uVs 11:28:43 I2X Average and RMS of Measurements VT = -37.729 +- 7.930 uVs I2X = 3.471 +- 0.730 uTm^2 11:28:45 Wire Position: X0 = 0.0000 mm, Y0 = 1.0000 mm 11:28:48 Integrated Voltage Measurement... 11:28:57 Voltage Offset Correction: v_off_ini = -123.848 uV, v_off_fin = -131.518 uV 11:28:57 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -214.643 uVs 11:28:57 Integrated Voltage Measurement... 11:29:06 Voltage Offset Correction: v_off_ini = -123.726 uV, v_off_fin = -116.134 uV 11:29:06 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 188.817 uVs 11:29:06 Integrated Voltage Measurement... 11:29:15 Voltage Offset Correction: v_off_ini = -118.981 uV, v_off_fin = -126.320 uV 11:29:15 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -195.090 uVs 11:29:15 Integrated Voltage Measurement... 11:29:24 Voltage Offset Correction: v_off_ini = -136.052 uV, v_off_fin = -128.307 uV 11:29:24 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 212.263 uVs 11:29:24 Integrated Voltage Measurement... 11:29:33 Voltage Offset Correction: v_off_ini = -132.267 uV, v_off_fin = -143.515 uV 11:29:33 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -208.999 uVs 11:29:33 Integrated Voltage Measurement... 11:29:42 Voltage Offset Correction: v_off_ini = -143.622 uV, v_off_fin = -136.513 uV 11:29:42 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 209.972 uVs 11:29:45 I2Y Average and RMS of Measurements VT = -204.964 +- 3.295 uVs I2Y = -18.857 +- 0.303 uTm^2 11:30:01 Wire Position: X0 = -2.0000 mm, Y0 = 0.0000 mm 11:30:05 Integrated Voltage Measurement... 11:30:14 Voltage Offset Correction: v_off_ini = -122.515 uV, v_off_fin = -111.643 uV 11:30:14 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -5.790 uVs 11:30:14 Integrated Voltage Measurement... 11:30:23 Voltage Offset Correction: v_off_ini = -122.621 uV, v_off_fin = -110.208 uV 11:30:23 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 9.857 uVs 11:30:23 Integrated Voltage Measurement... 11:30:32 Voltage Offset Correction: v_off_ini = -133.542 uV, v_off_fin = -141.435 uV 11:30:32 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 3.696 uVs 11:30:32 Integrated Voltage Measurement... 11:30:41 Voltage Offset Correction: v_off_ini = -149.821 uV, v_off_fin = -126.901 uV 11:30:41 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 19.467 uVs 11:30:41 Integrated Voltage Measurement... 11:30:50 Voltage Offset Correction: v_off_ini = -132.524 uV, v_off_fin = -141.264 uV 11:30:50 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -28.563 uVs 11:30:50 Integrated Voltage Measurement... 11:30:59 Voltage Offset Correction: v_off_ini = -133.992 uV, v_off_fin = -120.289 uV 11:30:59 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 6.577 uVs 11:31:03 I1X Average and RMS of Measurements VT = -11.093 +- 4.580 uVs I1X = 2.219 +- 0.916 uTm 11:31:06 Wire Position: X0 = -2.0000 mm, Y0 = 0.0000 mm 11:31:10 Integrated Voltage Measurement... 11:31:19 Voltage Offset Correction: v_off_ini = -134.676 uV, v_off_fin = -150.454 uV 11:31:19 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -132.703 uVs 11:31:19 Integrated Voltage Measurement... 11:31:28 Voltage Offset Correction: v_off_ini = -126.252 uV, v_off_fin = -123.337 uV 11:31:28 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.5000 mm dXY = -0.4999 mm VT = 133.605 uVs 11:31:28 Integrated Voltage Measurement... 11:31:37 Voltage Offset Correction: v_off_ini = -127.005 uV, v_off_fin = -135.638 uV 11:31:37 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.4999 mm VT = -144.512 uVs 11:31:37 Integrated Voltage Measurement... 11:31:46 Voltage Offset Correction: v_off_ini = -152.171 uV, v_off_fin = -126.320 uV 11:31:46 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 154.303 uVs 11:31:46 Integrated Voltage Measurement... 11:31:55 Voltage Offset Correction: v_off_ini = -132.145 uV, v_off_fin = -142.746 uV 11:31:55 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -164.991 uVs 11:31:55 Integrated Voltage Measurement... 11:32:04 Voltage Offset Correction: v_off_ini = -152.394 uV, v_off_fin = -143.022 uV 11:32:04 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.4999 mm dXY = -0.4999 mm VT = 150.234 uVs 11:32:08 I1Y Average and RMS of Measurements VT = -146.725 +- 10.164 uVs I1Y = -29.347 +- 2.033 uTm 11:32:13 Wire Position: X0 = -2.0000 mm, Y0 = 0.0000 mm 11:32:16 Integrated Voltage Measurement... 11:32:25 Voltage Offset Correction: v_off_ini = -126.255 uV, v_off_fin = -133.755 uV 11:32:25 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -29.258 uVs 11:32:25 Integrated Voltage Measurement... 11:32:34 Voltage Offset Correction: v_off_ini = -126.009 uV, v_off_fin = -141.867 uV 11:32:34 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 15.357 uVs 11:32:34 Integrated Voltage Measurement... 11:32:43 Voltage Offset Correction: v_off_ini = -142.136 uV, v_off_fin = -134.323 uV 11:32:43 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -10.377 uVs 11:32:43 Integrated Voltage Measurement... 11:32:52 Voltage Offset Correction: v_off_ini = -138.737 uV, v_off_fin = -136.888 uV 11:32:52 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 26.347 uVs 11:32:52 Integrated Voltage Measurement... 11:33:01 Voltage Offset Correction: v_off_ini = -129.203 uV, v_off_fin = -133.815 uV 11:33:01 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -23.037 uVs 11:33:01 Integrated Voltage Measurement... 11:33:10 Voltage Offset Correction: v_off_ini = -126.794 uV, v_off_fin = -134.467 uV 11:33:10 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 16.984 uVs 11:33:13 I2X Average and RMS of Measurements VT = -20.227 +- 1.618 uVs I2X = 1.861 +- 0.149 uTm^2 11:33:18 Wire Position: X0 = -2.0000 mm, Y0 = 0.0000 mm 11:33:20 Integrated Voltage Measurement... 11:33:29 Voltage Offset Correction: v_off_ini = -132.214 uV, v_off_fin = -124.353 uV 11:33:29 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -168.834 uVs 11:33:29 Integrated Voltage Measurement... 11:33:38 Voltage Offset Correction: v_off_ini = -130.135 uV, v_off_fin = -133.556 uV 11:33:38 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 183.281 uVs 11:33:38 Integrated Voltage Measurement... 11:33:47 Voltage Offset Correction: v_off_ini = -125.625 uV, v_off_fin = -134.938 uV 11:33:47 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -185.543 uVs 11:33:47 Integrated Voltage Measurement... 11:33:56 Voltage Offset Correction: v_off_ini = -142.321 uV, v_off_fin = -131.939 uV 11:33:56 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 165.426 uVs 11:33:56 Integrated Voltage Measurement... 11:34:05 Voltage Offset Correction: v_off_ini = -144.197 uV, v_off_fin = -145.331 uV 11:34:05 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -165.353 uVs 11:34:05 Integrated Voltage Measurement... 11:34:14 Voltage Offset Correction: v_off_ini = -142.284 uV, v_off_fin = -126.278 uV 11:34:14 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 172.472 uVs 11:34:17 I2Y Average and RMS of Measurements VT = -173.485 +- 3.241 uVs I2Y = -15.961 +- 0.298 uTm^2 11:34:21 Wire Position: X0 = -1.0000 mm, Y0 = 0.0000 mm 11:34:26 Integrated Voltage Measurement... 11:34:35 Voltage Offset Correction: v_off_ini = -134.338 uV, v_off_fin = -145.646 uV 11:34:35 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -29.042 uVs 11:34:35 Integrated Voltage Measurement... 11:34:44 Voltage Offset Correction: v_off_ini = -148.659 uV, v_off_fin = -134.532 uV 11:34:44 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 12.956 uVs 11:34:44 Integrated Voltage Measurement... 11:34:53 Voltage Offset Correction: v_off_ini = -135.476 uV, v_off_fin = -136.905 uV 11:34:53 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -33.306 uVs 11:34:53 Integrated Voltage Measurement... 11:35:02 Voltage Offset Correction: v_off_ini = -128.631 uV, v_off_fin = -137.163 uV 11:35:02 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 25.060 uVs 11:35:02 Integrated Voltage Measurement... 11:35:11 Voltage Offset Correction: v_off_ini = -128.414 uV, v_off_fin = -131.201 uV 11:35:11 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -25.442 uVs 11:35:11 Integrated Voltage Measurement... 11:35:20 Voltage Offset Correction: v_off_ini = -131.759 uV, v_off_fin = -116.132 uV 11:35:20 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 19.949 uVs 11:35:24 I1X Average and RMS of Measurements VT = -24.292 +- 3.527 uVs I1X = 4.858 +- 0.705 uTm 11:35:27 Wire Position: X0 = -1.0000 mm, Y0 = 0.0000 mm 11:35:31 Integrated Voltage Measurement... 11:35:40 Voltage Offset Correction: v_off_ini = -125.217 uV, v_off_fin = -129.673 uV 11:35:40 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -144.296 uVs 11:35:40 Integrated Voltage Measurement... 11:35:49 Voltage Offset Correction: v_off_ini = -129.051 uV, v_off_fin = -120.158 uV 11:35:49 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 136.870 uVs 11:35:49 Integrated Voltage Measurement... 11:35:58 Voltage Offset Correction: v_off_ini = -130.862 uV, v_off_fin = -112.684 uV 11:35:58 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -162.502 uVs 11:35:58 Integrated Voltage Measurement... 11:36:07 Voltage Offset Correction: v_off_ini = -115.981 uV, v_off_fin = -116.896 uV 11:36:07 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 142.276 uVs 11:36:07 Integrated Voltage Measurement... 11:36:16 Voltage Offset Correction: v_off_ini = -142.263 uV, v_off_fin = -140.748 uV 11:36:16 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -139.489 uVs 11:36:16 Integrated Voltage Measurement... 11:36:25 Voltage Offset Correction: v_off_ini = -120.696 uV, v_off_fin = -105.320 uV 11:36:25 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 157.416 uVs 11:36:29 I1Y Average and RMS of Measurements VT = -147.142 +- 4.908 uVs I1Y = -29.429 +- 0.982 uTm 11:36:34 Wire Position: X0 = -1.0000 mm, Y0 = 0.0000 mm 11:36:37 Integrated Voltage Measurement... 11:36:46 Voltage Offset Correction: v_off_ini = -111.853 uV, v_off_fin = -130.260 uV 11:36:46 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -48.757 uVs 11:36:46 Integrated Voltage Measurement... 11:36:55 Voltage Offset Correction: v_off_ini = -125.906 uV, v_off_fin = -135.358 uV 11:36:55 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 37.008 uVs 11:36:55 Integrated Voltage Measurement... 11:37:04 Voltage Offset Correction: v_off_ini = -115.052 uV, v_off_fin = -107.283 uV 11:37:04 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -23.705 uVs 11:37:04 Integrated Voltage Measurement... 11:37:13 Voltage Offset Correction: v_off_ini = -126.248 uV, v_off_fin = -117.845 uV 11:37:13 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 31.292 uVs 11:37:13 Integrated Voltage Measurement... 11:37:22 Voltage Offset Correction: v_off_ini = -127.070 uV, v_off_fin = -119.678 uV 11:37:22 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -18.893 uVs 11:37:22 Integrated Voltage Measurement... 11:37:31 Voltage Offset Correction: v_off_ini = -132.651 uV, v_off_fin = -130.780 uV 11:37:31 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 31.543 uVs 11:37:33 I2X Average and RMS of Measurements VT = -31.867 +- 7.845 uVs I2X = 2.932 +- 0.722 uTm^2 11:37:38 Wire Position: X0 = -1.0000 mm, Y0 = 0.0000 mm 11:37:41 Integrated Voltage Measurement... 11:37:50 Voltage Offset Correction: v_off_ini = -124.492 uV, v_off_fin = -129.705 uV 11:37:50 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -168.985 uVs 11:37:50 Integrated Voltage Measurement... 11:37:59 Voltage Offset Correction: v_off_ini = -132.686 uV, v_off_fin = -126.012 uV 11:37:59 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 174.685 uVs 11:37:59 Integrated Voltage Measurement... 11:38:08 Voltage Offset Correction: v_off_ini = -130.664 uV, v_off_fin = -122.097 uV 11:38:08 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -168.873 uVs 11:38:08 Integrated Voltage Measurement... 11:38:17 Voltage Offset Correction: v_off_ini = -132.645 uV, v_off_fin = -134.856 uV 11:38:17 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 169.928 uVs 11:38:17 Integrated Voltage Measurement... 11:38:26 Voltage Offset Correction: v_off_ini = -135.276 uV, v_off_fin = -133.385 uV 11:38:26 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -171.196 uVs 11:38:26 Integrated Voltage Measurement... 11:38:35 Voltage Offset Correction: v_off_ini = -118.605 uV, v_off_fin = -133.298 uV 11:38:35 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 165.459 uVs 11:38:37 I2Y Average and RMS of Measurements VT = -169.854 +- 1.468 uVs I2Y = -15.627 +- 0.135 uTm^2 11:38:42 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:38:46 Integrated Voltage Measurement... 11:38:55 Voltage Offset Correction: v_off_ini = -127.695 uV, v_off_fin = -139.338 uV 11:38:55 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -25.001 uVs 11:38:55 Integrated Voltage Measurement... 11:39:04 Voltage Offset Correction: v_off_ini = -140.273 uV, v_off_fin = -128.999 uV 11:39:04 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 38.029 uVs 11:39:04 Integrated Voltage Measurement... 11:39:13 Voltage Offset Correction: v_off_ini = -117.450 uV, v_off_fin = -130.723 uV 11:39:13 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -42.482 uVs 11:39:13 Integrated Voltage Measurement... 11:39:22 Voltage Offset Correction: v_off_ini = -122.464 uV, v_off_fin = -113.938 uV 11:39:22 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 25.626 uVs 11:39:22 Integrated Voltage Measurement... 11:39:31 Voltage Offset Correction: v_off_ini = -126.625 uV, v_off_fin = -108.321 uV 11:39:31 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -20.052 uVs 11:39:31 Integrated Voltage Measurement... 11:39:40 Voltage Offset Correction: v_off_ini = -107.035 uV, v_off_fin = -125.207 uV 11:39:40 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 42.570 uVs 11:39:45 I1X Average and RMS of Measurements VT = -32.293 +- 1.248 uVs I1X = 6.459 +- 0.250 uTm 11:39:47 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:39:51 Integrated Voltage Measurement... 11:40:00 Voltage Offset Correction: v_off_ini = -133.322 uV, v_off_fin = -133.610 uV 11:40:00 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.4999 mm dXY = 0.4999 mm VT = -143.675 uVs 11:40:00 Integrated Voltage Measurement... 11:40:09 Voltage Offset Correction: v_off_ini = -137.730 uV, v_off_fin = -128.615 uV 11:40:09 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 156.397 uVs 11:40:09 Integrated Voltage Measurement... 11:40:18 Voltage Offset Correction: v_off_ini = -118.878 uV, v_off_fin = -121.024 uV 11:40:18 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -160.408 uVs 11:40:18 Integrated Voltage Measurement... 11:40:27 Voltage Offset Correction: v_off_ini = -127.194 uV, v_off_fin = -122.218 uV 11:40:27 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.4999 mm dXY = -0.4999 mm VT = 158.922 uVs 11:40:27 Integrated Voltage Measurement... 11:40:36 Voltage Offset Correction: v_off_ini = -134.530 uV, v_off_fin = -146.275 uV 11:40:36 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.4999 mm dXY = 0.4999 mm VT = -146.549 uVs 11:40:36 Integrated Voltage Measurement... 11:40:45 Voltage Offset Correction: v_off_ini = -118.128 uV, v_off_fin = -122.208 uV 11:40:45 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 149.060 uVs 11:40:49 I1Y Average and RMS of Measurements VT = -152.502 +- 5.146 uVs I1Y = -30.503 +- 1.029 uTm 11:40:54 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:40:57 Integrated Voltage Measurement... 11:41:06 Voltage Offset Correction: v_off_ini = -143.172 uV, v_off_fin = -128.570 uV 11:41:06 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -47.684 uVs 11:41:06 Integrated Voltage Measurement... 11:41:15 Voltage Offset Correction: v_off_ini = -137.639 uV, v_off_fin = -127.276 uV 11:41:15 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 39.637 uVs 11:41:15 Integrated Voltage Measurement... 11:41:24 Voltage Offset Correction: v_off_ini = -113.249 uV, v_off_fin = -134.371 uV 11:41:24 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -21.354 uVs 11:41:24 Integrated Voltage Measurement... 11:41:33 Voltage Offset Correction: v_off_ini = -131.864 uV, v_off_fin = -119.072 uV 11:41:33 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 44.756 uVs 11:41:33 Integrated Voltage Measurement... 11:41:42 Voltage Offset Correction: v_off_ini = -131.321 uV, v_off_fin = -118.180 uV 11:41:42 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -29.842 uVs 11:41:42 Integrated Voltage Measurement... 11:41:51 Voltage Offset Correction: v_off_ini = -124.575 uV, v_off_fin = -140.962 uV 11:41:51 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 40.046 uVs 11:41:53 I2X Average and RMS of Measurements VT = -37.220 +- 4.619 uVs I2X = 3.424 +- 0.425 uTm^2 11:41:59 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:42:01 Integrated Voltage Measurement... 11:42:10 Voltage Offset Correction: v_off_ini = -117.056 uV, v_off_fin = -113.141 uV 11:42:10 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -161.785 uVs 11:42:10 Integrated Voltage Measurement... 11:42:19 Voltage Offset Correction: v_off_ini = -112.554 uV, v_off_fin = -124.392 uV 11:42:19 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 171.006 uVs 11:42:19 Integrated Voltage Measurement... 11:42:28 Voltage Offset Correction: v_off_ini = -110.291 uV, v_off_fin = -125.082 uV 11:42:28 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -193.864 uVs 11:42:28 Integrated Voltage Measurement... 11:42:37 Voltage Offset Correction: v_off_ini = -128.688 uV, v_off_fin = -127.878 uV 11:42:37 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 186.047 uVs 11:42:37 Integrated Voltage Measurement... 11:42:46 Voltage Offset Correction: v_off_ini = -128.176 uV, v_off_fin = -125.578 uV 11:42:46 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -189.435 uVs 11:42:46 Integrated Voltage Measurement... 11:42:55 Voltage Offset Correction: v_off_ini = -131.762 uV, v_off_fin = -130.802 uV 11:42:55 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 190.119 uVs 11:42:58 I2Y Average and RMS of Measurements VT = -182.043 +- 11.065 uVs I2Y = -16.748 +- 1.018 uTm^2 11:43:02 Wire Position: X0 = 1.0000 mm, Y0 = 0.0000 mm 11:43:06 Integrated Voltage Measurement... 11:43:15 Voltage Offset Correction: v_off_ini = -135.378 uV, v_off_fin = -122.025 uV 11:43:15 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -35.536 uVs 11:43:15 Integrated Voltage Measurement... 11:43:24 Voltage Offset Correction: v_off_ini = -125.518 uV, v_off_fin = -118.574 uV 11:43:24 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 59.954 uVs 11:43:24 Integrated Voltage Measurement... 11:43:33 Voltage Offset Correction: v_off_ini = -120.844 uV, v_off_fin = -126.177 uV 11:43:33 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -42.254 uVs 11:43:33 Integrated Voltage Measurement... 11:43:42 Voltage Offset Correction: v_off_ini = -117.571 uV, v_off_fin = -108.566 uV 11:43:42 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 28.528 uVs 11:43:42 Integrated Voltage Measurement... 11:43:51 Voltage Offset Correction: v_off_ini = -112.205 uV, v_off_fin = -128.338 uV 11:43:51 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -49.870 uVs 11:43:51 Integrated Voltage Measurement... 11:44:00 Voltage Offset Correction: v_off_ini = -126.795 uV, v_off_fin = -128.088 uV 11:44:00 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 43.677 uVs 11:44:04 I1X Average and RMS of Measurements VT = -43.303 +- 5.609 uVs I1X = 8.661 +- 1.122 uTm 11:44:07 Wire Position: X0 = 1.0000 mm, Y0 = 0.0000 mm 11:44:11 Integrated Voltage Measurement... 11:44:20 Voltage Offset Correction: v_off_ini = -120.085 uV, v_off_fin = -143.226 uV 11:44:20 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -146.396 uVs 11:44:20 Integrated Voltage Measurement... 11:44:29 Voltage Offset Correction: v_off_ini = -143.042 uV, v_off_fin = -108.973 uV 11:44:29 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 158.204 uVs 11:44:29 Integrated Voltage Measurement... 11:44:38 Voltage Offset Correction: v_off_ini = -122.668 uV, v_off_fin = -132.959 uV 11:44:38 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -153.248 uVs 11:44:38 Integrated Voltage Measurement... 11:44:47 Voltage Offset Correction: v_off_ini = -126.777 uV, v_off_fin = -117.689 uV 11:44:47 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 162.015 uVs 11:44:47 Integrated Voltage Measurement... 11:44:56 Voltage Offset Correction: v_off_ini = -112.975 uV, v_off_fin = -125.965 uV 11:44:56 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -156.218 uVs 11:44:56 Integrated Voltage Measurement... 11:45:05 Voltage Offset Correction: v_off_ini = -122.106 uV, v_off_fin = -118.730 uV 11:45:05 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 160.250 uVs 11:45:09 I1Y Average and RMS of Measurements VT = -156.055 +- 2.667 uVs I1Y = -31.211 +- 0.533 uTm 11:45:14 Wire Position: X0 = 1.0000 mm, Y0 = 0.0000 mm 11:45:17 Integrated Voltage Measurement... 11:45:26 Voltage Offset Correction: v_off_ini = -115.977 uV, v_off_fin = -117.417 uV 11:45:26 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -60.841 uVs 11:45:26 Integrated Voltage Measurement... 11:45:35 Voltage Offset Correction: v_off_ini = -128.659 uV, v_off_fin = -119.511 uV 11:45:35 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 60.312 uVs 11:45:35 Integrated Voltage Measurement... 11:45:44 Voltage Offset Correction: v_off_ini = -120.936 uV, v_off_fin = -127.720 uV 11:45:44 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -45.271 uVs 11:45:44 Integrated Voltage Measurement... 11:45:53 Voltage Offset Correction: v_off_ini = -113.852 uV, v_off_fin = -112.537 uV 11:45:53 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 41.024 uVs 11:45:53 Integrated Voltage Measurement... 11:46:02 Voltage Offset Correction: v_off_ini = -113.462 uV, v_off_fin = -115.736 uV 11:46:02 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -69.627 uVs 11:46:02 Integrated Voltage Measurement... 11:46:11 Voltage Offset Correction: v_off_ini = -115.596 uV, v_off_fin = -135.766 uV 11:46:11 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 55.634 uVs 11:46:13 I2X Average and RMS of Measurements VT = -55.451 +- 8.741 uVs I2X = 5.102 +- 0.804 uTm^2 11:46:18 Wire Position: X0 = 1.0000 mm, Y0 = 0.0000 mm 11:46:21 Integrated Voltage Measurement... 11:46:30 Voltage Offset Correction: v_off_ini = -121.331 uV, v_off_fin = -119.088 uV 11:46:30 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -169.898 uVs 11:46:30 Integrated Voltage Measurement... 11:46:39 Voltage Offset Correction: v_off_ini = -110.317 uV, v_off_fin = -107.892 uV 11:46:39 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = 177.173 uVs 11:46:39 Integrated Voltage Measurement... 11:46:48 Voltage Offset Correction: v_off_ini = -118.583 uV, v_off_fin = -120.506 uV 11:46:48 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = -164.917 uVs 11:46:48 Integrated Voltage Measurement... 11:46:57 Voltage Offset Correction: v_off_ini = -92.960 uV, v_off_fin = -91.363 uV 11:46:57 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 165.535 uVs 11:46:57 Integrated Voltage Measurement... 11:47:06 Voltage Offset Correction: v_off_ini = -98.069 uV, v_off_fin = -117.474 uV 11:47:06 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = -169.883 uVs 11:47:06 Integrated Voltage Measurement... 11:47:15 Voltage Offset Correction: v_off_ini = -106.310 uV, v_off_fin = -94.200 uV 11:47:15 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 158.780 uVs 11:47:17 I2Y Average and RMS of Measurements VT = -167.698 +- 4.144 uVs I2Y = -15.429 +- 0.381 uTm^2 11:47:22 Wire Position: X0 = 2.0000 mm, Y0 = 0.0000 mm 11:47:26 Integrated Voltage Measurement... 11:47:35 Voltage Offset Correction: v_off_ini = -98.612 uV, v_off_fin = -96.248 uV 11:47:35 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -39.734 uVs 11:47:35 Integrated Voltage Measurement... 11:47:44 Voltage Offset Correction: v_off_ini = -89.276 uV, v_off_fin = -88.848 uV 11:47:44 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 71.635 uVs 11:47:44 Integrated Voltage Measurement... 11:47:53 Voltage Offset Correction: v_off_ini = -97.977 uV, v_off_fin = -87.743 uV 11:47:53 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -78.739 uVs 11:47:53 Integrated Voltage Measurement... 11:48:02 Voltage Offset Correction: v_off_ini = -94.614 uV, v_off_fin = -91.461 uV 11:48:02 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 55.942 uVs 11:48:02 Integrated Voltage Measurement... 11:48:11 Voltage Offset Correction: v_off_ini = -83.865 uV, v_off_fin = -83.890 uV 11:48:11 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -40.993 uVs 11:48:11 Integrated Voltage Measurement... 11:48:20 Voltage Offset Correction: v_off_ini = -77.994 uV, v_off_fin = -81.163 uV 11:48:20 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 58.499 uVs 11:48:25 I1X Average and RMS of Measurements VT = -57.590 +- 7.308 uVs I1X = 11.518 +- 1.462 uTm 11:48:27 Wire Position: X0 = 2.0000 mm, Y0 = 0.0000 mm 11:48:31 Integrated Voltage Measurement... 11:48:40 Voltage Offset Correction: v_off_ini = -80.773 uV, v_off_fin = -85.553 uV 11:48:40 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.4999 mm dXY = 0.4999 mm VT = -160.702 uVs 11:48:40 Integrated Voltage Measurement... 11:48:49 Voltage Offset Correction: v_off_ini = -77.741 uV, v_off_fin = -80.677 uV 11:48:49 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 167.361 uVs 11:48:49 Integrated Voltage Measurement... 11:48:58 Voltage Offset Correction: v_off_ini = -76.950 uV, v_off_fin = -93.129 uV 11:48:58 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -167.296 uVs 11:48:58 Integrated Voltage Measurement... 11:49:07 Voltage Offset Correction: v_off_ini = -80.161 uV, v_off_fin = -75.547 uV 11:49:07 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 152.885 uVs 11:49:07 Integrated Voltage Measurement... 11:49:16 Voltage Offset Correction: v_off_ini = -78.599 uV, v_off_fin = -75.480 uV 11:49:16 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -163.579 uVs 11:49:16 Integrated Voltage Measurement... 11:49:25 Voltage Offset Correction: v_off_ini = -83.298 uV, v_off_fin = -71.206 uV 11:49:25 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 161.683 uVs 11:49:29 I1Y Average and RMS of Measurements VT = -162.251 +- 1.631 uVs I1Y = -32.451 +- 0.327 uTm 11:49:34 Wire Position: X0 = 2.0000 mm, Y0 = 0.0000 mm 11:49:37 Integrated Voltage Measurement... 11:49:46 Voltage Offset Correction: v_off_ini = -61.705 uV, v_off_fin = -61.392 uV 11:49:46 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -90.379 uVs 11:49:46 Integrated Voltage Measurement... 11:49:55 Voltage Offset Correction: v_off_ini = -66.153 uV, v_off_fin = -85.133 uV 11:49:55 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 55.092 uVs 11:49:55 Integrated Voltage Measurement... 11:50:04 Voltage Offset Correction: v_off_ini = -83.457 uV, v_off_fin = -75.037 uV 11:50:04 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -76.960 uVs 11:50:04 Integrated Voltage Measurement... 11:50:13 Voltage Offset Correction: v_off_ini = -79.182 uV, v_off_fin = -77.900 uV 11:50:13 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 72.918 uVs 11:50:13 Integrated Voltage Measurement... 11:50:22 Voltage Offset Correction: v_off_ini = -78.914 uV, v_off_fin = -66.256 uV 11:50:22 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -65.212 uVs 11:50:22 Integrated Voltage Measurement... 11:50:31 Voltage Offset Correction: v_off_ini = -76.226 uV, v_off_fin = -87.226 uV 11:50:31 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 53.566 uVs 11:50:33 I2X Average and RMS of Measurements VT = -69.021 +- 6.870 uVs I2X = 6.350 +- 0.632 uTm^2 11:50:39 Wire Position: X0 = 2.0000 mm, Y0 = 0.0000 mm 11:50:41 Integrated Voltage Measurement... 11:50:50 Voltage Offset Correction: v_off_ini = -83.068 uV, v_off_fin = -83.672 uV 11:50:50 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -175.924 uVs 11:50:50 Integrated Voltage Measurement... 11:50:59 Voltage Offset Correction: v_off_ini = -83.350 uV, v_off_fin = -79.589 uV 11:50:59 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 201.272 uVs 11:50:59 Integrated Voltage Measurement... 11:51:08 Voltage Offset Correction: v_off_ini = -86.176 uV, v_off_fin = -77.630 uV 11:51:08 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -160.844 uVs 11:51:08 Integrated Voltage Measurement... 11:51:17 Voltage Offset Correction: v_off_ini = -85.893 uV, v_off_fin = -88.361 uV 11:51:17 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 153.606 uVs 11:51:17 Integrated Voltage Measurement... 11:51:26 Voltage Offset Correction: v_off_ini = -92.555 uV, v_off_fin = -93.897 uV 11:51:26 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -185.385 uVs 11:51:26 Integrated Voltage Measurement... 11:51:35 Voltage Offset Correction: v_off_ini = -79.380 uV, v_off_fin = -66.736 uV 11:51:35 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 168.083 uVs 11:51:38 I2Y Average and RMS of Measurements VT = -174.186 +- 12.934 uVs I2Y = -16.025 +- 1.190 uTm^2 11:51:45 Wire Position: X0 = 0.0000 mm, Y0 = -1.0000 mm 11:51:50 Integrated Voltage Measurement... 11:51:59 Voltage Offset Correction: v_off_ini = -89.376 uV, v_off_fin = -79.230 uV 11:51:59 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -39.425 uVs 11:51:59 Integrated Voltage Measurement... 11:52:08 Voltage Offset Correction: v_off_ini = -72.342 uV, v_off_fin = -106.587 uV 11:52:08 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 26.639 uVs 11:52:08 Integrated Voltage Measurement... 11:52:17 Voltage Offset Correction: v_off_ini = -93.436 uV, v_off_fin = -87.152 uV 11:52:17 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -31.904 uVs 11:52:17 Integrated Voltage Measurement... 11:52:26 Voltage Offset Correction: v_off_ini = -85.047 uV, v_off_fin = -72.789 uV 11:52:26 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 46.028 uVs 11:52:26 Integrated Voltage Measurement... 11:52:35 Voltage Offset Correction: v_off_ini = -85.736 uV, v_off_fin = -80.503 uV 11:52:35 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -42.839 uVs 11:52:35 Integrated Voltage Measurement... 11:52:44 Voltage Offset Correction: v_off_ini = -84.926 uV, v_off_fin = -90.110 uV 11:52:44 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 20.823 uVs 11:52:48 I1X Average and RMS of Measurements VT = -34.610 +- 3.119 uVs I1X = 6.922 +- 0.624 uTm 11:52:50 Wire Position: X0 = 0.0000 mm, Y0 = -1.0000 mm 11:52:55 Integrated Voltage Measurement... 11:53:04 Voltage Offset Correction: v_off_ini = -84.966 uV, v_off_fin = -91.595 uV 11:53:04 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -139.867 uVs 11:53:04 Integrated Voltage Measurement... 11:53:13 Voltage Offset Correction: v_off_ini = -88.560 uV, v_off_fin = -83.216 uV 11:53:13 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 151.528 uVs 11:53:13 Integrated Voltage Measurement... 11:53:22 Voltage Offset Correction: v_off_ini = -72.229 uV, v_off_fin = -90.845 uV 11:53:22 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -157.596 uVs 11:53:22 Integrated Voltage Measurement... 11:53:31 Voltage Offset Correction: v_off_ini = -72.119 uV, v_off_fin = -80.510 uV 11:53:31 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 156.875 uVs 11:53:31 Integrated Voltage Measurement... 11:53:40 Voltage Offset Correction: v_off_ini = -87.497 uV, v_off_fin = -98.608 uV 11:53:40 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -151.371 uVs 11:53:40 Integrated Voltage Measurement... 11:53:48 Voltage Offset Correction: v_off_ini = -72.634 uV, v_off_fin = -68.798 uV 11:53:49 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 150.972 uVs 11:53:53 I1Y Average and RMS of Measurements VT = -151.368 +- 4.712 uVs I1Y = -30.274 +- 0.942 uTm 11:53:55 Wire Position: X0 = 0.0000 mm, Y0 = -1.0000 mm 11:53:58 Integrated Voltage Measurement... 11:54:07 Voltage Offset Correction: v_off_ini = -93.037 uV, v_off_fin = -84.286 uV 11:54:07 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -26.690 uVs 11:54:07 Integrated Voltage Measurement... 11:54:16 Voltage Offset Correction: v_off_ini = -96.007 uV, v_off_fin = -90.201 uV 11:54:16 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 34.953 uVs 11:54:16 Integrated Voltage Measurement... 11:54:25 Voltage Offset Correction: v_off_ini = -79.594 uV, v_off_fin = -81.498 uV 11:54:25 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -33.284 uVs 11:54:25 Integrated Voltage Measurement... 11:54:34 Voltage Offset Correction: v_off_ini = -68.529 uV, v_off_fin = -101.574 uV 11:54:34 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 43.362 uVs 11:54:34 Integrated Voltage Measurement... 11:54:43 Voltage Offset Correction: v_off_ini = -102.202 uV, v_off_fin = -82.555 uV 11:54:43 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -27.156 uVs 11:54:43 Integrated Voltage Measurement... 11:54:52 Voltage Offset Correction: v_off_ini = -91.620 uV, v_off_fin = -93.180 uV 11:54:52 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 59.720 uVs 11:54:54 I2X Average and RMS of Measurements VT = -37.528 +- 5.181 uVs I2X = 3.453 +- 0.477 uTm^2 11:54:57 Wire Position: X0 = 0.0000 mm, Y0 = -1.0000 mm 11:54:59 Integrated Voltage Measurement... 11:55:08 Voltage Offset Correction: v_off_ini = -103.666 uV, v_off_fin = -95.736 uV 11:55:08 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -163.648 uVs 11:55:08 Integrated Voltage Measurement... 11:55:17 Voltage Offset Correction: v_off_ini = -83.308 uV, v_off_fin = -75.791 uV 11:55:17 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 177.209 uVs 11:55:18 Integrated Voltage Measurement... 11:55:26 Voltage Offset Correction: v_off_ini = -84.264 uV, v_off_fin = -90.795 uV 11:55:26 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -154.503 uVs 11:55:27 Integrated Voltage Measurement... 11:55:35 Voltage Offset Correction: v_off_ini = -100.820 uV, v_off_fin = -104.668 uV 11:55:36 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 181.411 uVs 11:55:36 Integrated Voltage Measurement... 11:55:45 Voltage Offset Correction: v_off_ini = -97.883 uV, v_off_fin = -92.911 uV 11:55:45 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -160.560 uVs 11:55:45 Integrated Voltage Measurement... 11:55:54 Voltage Offset Correction: v_off_ini = -102.368 uV, v_off_fin = -96.914 uV 11:55:54 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 187.746 uVs 11:55:56 I2Y Average and RMS of Measurements VT = -170.846 +- 2.547 uVs I2Y = -15.718 +- 0.234 uTm^2 11:56:00 Wire Position: X0 = 0.0000 mm, Y0 = -0.5000 mm 11:56:04 Integrated Voltage Measurement... 11:56:13 Voltage Offset Correction: v_off_ini = -101.670 uV, v_off_fin = -113.451 uV 11:56:13 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -25.736 uVs 11:56:13 Integrated Voltage Measurement... 11:56:22 Voltage Offset Correction: v_off_ini = -109.664 uV, v_off_fin = -98.638 uV 11:56:22 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 58.259 uVs 11:56:22 Integrated Voltage Measurement... 11:56:31 Voltage Offset Correction: v_off_ini = -93.038 uV, v_off_fin = -94.620 uV 11:56:31 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -44.209 uVs 11:56:31 Integrated Voltage Measurement... 11:56:40 Voltage Offset Correction: v_off_ini = -89.342 uV, v_off_fin = -101.091 uV 11:56:40 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 49.880 uVs 11:56:40 Integrated Voltage Measurement... 11:56:49 Voltage Offset Correction: v_off_ini = -100.729 uV, v_off_fin = -120.996 uV 11:56:49 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -34.326 uVs 11:56:49 Integrated Voltage Measurement... 11:56:58 Voltage Offset Correction: v_off_ini = -110.557 uV, v_off_fin = -104.327 uV 11:56:58 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 23.550 uVs 11:57:02 I1X Average and RMS of Measurements VT = -39.327 +- 7.629 uVs I1X = 7.865 +- 1.526 uTm 11:57:05 Wire Position: X0 = 0.0000 mm, Y0 = -0.5000 mm 11:57:09 Integrated Voltage Measurement... 11:57:18 Voltage Offset Correction: v_off_ini = -99.972 uV, v_off_fin = -107.869 uV 11:57:18 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -147.795 uVs 11:57:18 Integrated Voltage Measurement... 11:57:27 Voltage Offset Correction: v_off_ini = -110.437 uV, v_off_fin = -92.483 uV 11:57:27 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 151.388 uVs 11:57:27 Integrated Voltage Measurement... 11:57:36 Voltage Offset Correction: v_off_ini = -97.859 uV, v_off_fin = -117.018 uV 11:57:36 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -157.578 uVs 11:57:36 Integrated Voltage Measurement... 11:57:45 Voltage Offset Correction: v_off_ini = -119.577 uV, v_off_fin = -97.080 uV 11:57:45 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 164.527 uVs 11:57:45 Integrated Voltage Measurement... 11:57:54 Voltage Offset Correction: v_off_ini = -116.477 uV, v_off_fin = -107.640 uV 11:57:54 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -150.867 uVs 11:57:54 Integrated Voltage Measurement... 11:58:03 Voltage Offset Correction: v_off_ini = -97.605 uV, v_off_fin = -98.373 uV 11:58:03 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 157.670 uVs 11:58:07 I1Y Average and RMS of Measurements VT = -154.971 +- 4.705 uVs I1Y = -30.994 +- 0.941 uTm 11:58:10 Wire Position: X0 = 0.0000 mm, Y0 = -0.5000 mm 11:58:12 Integrated Voltage Measurement... 11:58:21 Voltage Offset Correction: v_off_ini = -103.413 uV, v_off_fin = -98.398 uV 11:58:21 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -31.531 uVs 11:58:22 Integrated Voltage Measurement... 11:58:30 Voltage Offset Correction: v_off_ini = -95.133 uV, v_off_fin = -99.697 uV 11:58:31 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 60.492 uVs 11:58:31 Integrated Voltage Measurement... 11:58:40 Voltage Offset Correction: v_off_ini = -89.287 uV, v_off_fin = -87.422 uV 11:58:40 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -45.924 uVs 11:58:40 Integrated Voltage Measurement... 11:58:49 Voltage Offset Correction: v_off_ini = -70.258 uV, v_off_fin = -79.814 uV 11:58:49 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 27.398 uVs 11:58:49 Integrated Voltage Measurement... 11:58:58 Voltage Offset Correction: v_off_ini = -75.384 uV, v_off_fin = -92.360 uV 11:58:58 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -22.105 uVs 11:58:58 Integrated Voltage Measurement... 11:59:07 Voltage Offset Correction: v_off_ini = -75.450 uV, v_off_fin = -83.157 uV 11:59:07 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 34.705 uVs 11:59:09 I2X Average and RMS of Measurements VT = -37.026 +- 7.193 uVs I2X = 3.406 +- 0.662 uTm^2 11:59:12 Wire Position: X0 = 0.0000 mm, Y0 = -0.5000 mm 11:59:14 Integrated Voltage Measurement... 11:59:23 Voltage Offset Correction: v_off_ini = -96.686 uV, v_off_fin = -94.796 uV 11:59:23 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -166.289 uVs 11:59:23 Integrated Voltage Measurement... 11:59:32 Voltage Offset Correction: v_off_ini = -89.178 uV, v_off_fin = -88.689 uV 11:59:32 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = 165.139 uVs 11:59:32 Integrated Voltage Measurement... 11:59:41 Voltage Offset Correction: v_off_ini = -79.079 uV, v_off_fin = -84.446 uV 11:59:41 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -158.096 uVs 11:59:41 Integrated Voltage Measurement... 11:59:50 Voltage Offset Correction: v_off_ini = -90.388 uV, v_off_fin = -88.218 uV 11:59:50 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 170.681 uVs 11:59:50 Integrated Voltage Measurement... 11:59:59 Voltage Offset Correction: v_off_ini = -87.300 uV, v_off_fin = -77.115 uV 11:59:59 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -160.785 uVs 11:59:59 Integrated Voltage Measurement... 12:00:08 Voltage Offset Correction: v_off_ini = -84.636 uV, v_off_fin = -70.657 uV 12:00:08 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 165.573 uVs 12:00:11 I2Y Average and RMS of Measurements VT = -164.427 +- 1.035 uVs I2Y = -15.128 +- 0.096 uTm^2 12:00:15 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 12:00:19 Integrated Voltage Measurement... 12:00:28 Voltage Offset Correction: v_off_ini = -66.564 uV, v_off_fin = -80.955 uV 12:00:28 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -26.180 uVs 12:00:28 Integrated Voltage Measurement... 12:00:37 Voltage Offset Correction: v_off_ini = -73.157 uV, v_off_fin = -73.420 uV 12:00:37 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 29.769 uVs 12:00:37 Integrated Voltage Measurement... 12:00:46 Voltage Offset Correction: v_off_ini = -74.765 uV, v_off_fin = -61.615 uV 12:00:46 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -41.437 uVs 12:00:46 Integrated Voltage Measurement... 12:00:55 Voltage Offset Correction: v_off_ini = -73.493 uV, v_off_fin = -97.447 uV 12:00:55 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 30.490 uVs 12:00:55 Integrated Voltage Measurement... 12:01:04 Voltage Offset Correction: v_off_ini = -92.246 uV, v_off_fin = -99.040 uV 12:01:04 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -35.054 uVs 12:01:04 Integrated Voltage Measurement... 12:01:13 Voltage Offset Correction: v_off_ini = -93.152 uV, v_off_fin = -95.971 uV 12:01:13 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 27.507 uVs 12:01:17 I1X Average and RMS of Measurements VT = -31.739 +- 3.278 uVs I1X = 6.348 +- 0.655 uTm 12:01:20 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 12:01:24 Integrated Voltage Measurement... 12:01:33 Voltage Offset Correction: v_off_ini = -99.307 uV, v_off_fin = -99.957 uV 12:01:33 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = -161.403 uVs 12:01:33 Integrated Voltage Measurement... 12:01:42 Voltage Offset Correction: v_off_ini = -98.153 uV, v_off_fin = -89.470 uV 12:01:42 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 154.639 uVs 12:01:42 Integrated Voltage Measurement... 12:01:51 Voltage Offset Correction: v_off_ini = -91.728 uV, v_off_fin = -110.344 uV 12:01:51 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -147.751 uVs 12:01:51 Integrated Voltage Measurement... 12:02:00 Voltage Offset Correction: v_off_ini = -95.002 uV, v_off_fin = -79.410 uV 12:02:00 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 152.887 uVs 12:02:00 Integrated Voltage Measurement... 12:02:09 Voltage Offset Correction: v_off_ini = -92.675 uV, v_off_fin = -96.327 uV 12:02:09 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -164.684 uVs 12:02:09 Integrated Voltage Measurement... 12:02:18 Voltage Offset Correction: v_off_ini = -87.125 uV, v_off_fin = -90.116 uV 12:02:18 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 155.959 uVs 12:02:22 I1Y Average and RMS of Measurements VT = -156.221 +- 4.277 uVs I1Y = -31.244 +- 0.856 uTm 12:02:25 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 12:02:27 Integrated Voltage Measurement... 12:02:36 Voltage Offset Correction: v_off_ini = -83.149 uV, v_off_fin = -79.328 uV 12:02:36 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -45.169 uVs 12:02:37 Integrated Voltage Measurement... 12:02:45 Voltage Offset Correction: v_off_ini = -92.239 uV, v_off_fin = -94.567 uV 12:02:46 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 47.380 uVs 12:02:46 Integrated Voltage Measurement... 12:02:55 Voltage Offset Correction: v_off_ini = -101.470 uV, v_off_fin = -95.396 uV 12:02:55 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -45.222 uVs 12:02:55 Integrated Voltage Measurement... 12:03:04 Voltage Offset Correction: v_off_ini = -100.481 uV, v_off_fin = -109.984 uV 12:03:04 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 40.555 uVs 12:03:04 Integrated Voltage Measurement... 12:03:13 Voltage Offset Correction: v_off_ini = -97.583 uV, v_off_fin = -87.556 uV 12:03:13 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -30.057 uVs 12:03:13 Integrated Voltage Measurement... 12:03:22 Voltage Offset Correction: v_off_ini = -83.281 uV, v_off_fin = -110.544 uV 12:03:22 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 40.678 uVs 12:03:24 I2X Average and RMS of Measurements VT = -41.510 +- 4.558 uVs I2X = 3.819 +- 0.419 uTm^2 12:03:27 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 12:03:29 Integrated Voltage Measurement... 12:03:38 Voltage Offset Correction: v_off_ini = -97.271 uV, v_off_fin = -104.724 uV 12:03:38 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -171.978 uVs 12:03:38 Integrated Voltage Measurement... 12:03:47 Voltage Offset Correction: v_off_ini = -110.061 uV, v_off_fin = -101.615 uV 12:03:47 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 162.345 uVs 12:03:47 Integrated Voltage Measurement... 12:03:56 Voltage Offset Correction: v_off_ini = -116.273 uV, v_off_fin = -104.334 uV 12:03:56 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -175.536 uVs 12:03:56 Integrated Voltage Measurement... 12:04:05 Voltage Offset Correction: v_off_ini = -109.200 uV, v_off_fin = -105.316 uV 12:04:05 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 183.180 uVs 12:04:05 Integrated Voltage Measurement... 12:04:14 Voltage Offset Correction: v_off_ini = -104.606 uV, v_off_fin = -105.544 uV 12:04:14 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -159.719 uVs 12:04:14 Integrated Voltage Measurement... 12:04:23 Voltage Offset Correction: v_off_ini = -106.993 uV, v_off_fin = -124.374 uV 12:04:23 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 163.851 uVs 12:04:26 I2Y Average and RMS of Measurements VT = -169.435 +- 7.352 uVs I2Y = -15.588 +- 0.676 uTm^2 12:04:30 Wire Position: X0 = 0.0000 mm, Y0 = 0.5000 mm 12:04:34 Integrated Voltage Measurement... 12:04:43 Voltage Offset Correction: v_off_ini = -107.648 uV, v_off_fin = -108.264 uV 12:04:43 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4998 mm dXY = -0.4999 mm VT = -41.241 uVs 12:04:43 Integrated Voltage Measurement... 12:04:52 Voltage Offset Correction: v_off_ini = -125.784 uV, v_off_fin = -129.048 uV 12:04:52 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 37.495 uVs 12:04:52 Integrated Voltage Measurement... 12:05:01 Voltage Offset Correction: v_off_ini = -132.011 uV, v_off_fin = -133.009 uV 12:05:01 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -9.982 uVs 12:05:01 Integrated Voltage Measurement... 12:05:10 Voltage Offset Correction: v_off_ini = -118.021 uV, v_off_fin = -119.434 uV 12:05:10 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 26.028 uVs 12:05:10 Integrated Voltage Measurement... 12:05:19 Voltage Offset Correction: v_off_ini = -119.509 uV, v_off_fin = -119.055 uV 12:05:19 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -28.326 uVs 12:05:19 Integrated Voltage Measurement... 12:05:28 Voltage Offset Correction: v_off_ini = -101.980 uV, v_off_fin = -121.288 uV 12:05:28 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 21.130 uVs 12:05:33 I1X Average and RMS of Measurements VT = -27.367 +- 8.919 uVs I1X = 5.474 +- 1.784 uTm 12:05:35 Wire Position: X0 = 0.0000 mm, Y0 = 0.5000 mm 12:05:39 Integrated Voltage Measurement... 12:05:48 Voltage Offset Correction: v_off_ini = -117.630 uV, v_off_fin = -118.315 uV 12:05:48 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = -153.736 uVs 12:05:48 Integrated Voltage Measurement... 12:05:57 Voltage Offset Correction: v_off_ini = -112.866 uV, v_off_fin = -101.647 uV 12:05:57 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 157.751 uVs 12:05:57 Integrated Voltage Measurement... 12:06:06 Voltage Offset Correction: v_off_ini = -106.662 uV, v_off_fin = -111.745 uV 12:06:06 Integrated Voltage Measurement dXYus = 0.5001 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -175.693 uVs 12:06:06 Integrated Voltage Measurement... 12:06:15 Voltage Offset Correction: v_off_ini = -106.794 uV, v_off_fin = -89.615 uV 12:06:15 Integrated Voltage Measurement dXYus = -0.5001 mm, dXYds = -0.5001 mm dXY = -0.5001 mm VT = 171.993 uVs 12:06:15 Integrated Voltage Measurement... 12:06:24 Voltage Offset Correction: v_off_ini = -103.802 uV, v_off_fin = -100.649 uV 12:06:24 Integrated Voltage Measurement dXYus = 0.5001 mm, dXYds = 0.5001 mm dXY = 0.5001 mm VT = -154.218 uVs 12:06:24 Integrated Voltage Measurement... 12:06:33 Voltage Offset Correction: v_off_ini = -93.811 uV, v_off_fin = -92.968 uV 12:06:33 Integrated Voltage Measurement dXYus = -0.5001 mm, dXYds = -0.5001 mm dXY = -0.5001 mm VT = 161.821 uVs 12:06:37 I1Y Average and RMS of Measurements VT = -162.535 +- 8.050 uVs I1Y = -32.504 +- 1.608 uTm 12:06:40 Wire Position: X0 = 0.0000 mm, Y0 = 0.5000 mm 12:06:43 Integrated Voltage Measurement... 12:06:52 Voltage Offset Correction: v_off_ini = -91.862 uV, v_off_fin = -98.811 uV 12:06:52 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -50.377 uVs 12:06:52 Integrated Voltage Measurement... 12:07:01 Voltage Offset Correction: v_off_ini = -101.916 uV, v_off_fin = -108.257 uV 12:07:01 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 35.564 uVs 12:07:01 Integrated Voltage Measurement... 12:07:10 Voltage Offset Correction: v_off_ini = -118.289 uV, v_off_fin = -100.457 uV 12:07:10 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -20.753 uVs 12:07:10 Integrated Voltage Measurement... 12:07:19 Voltage Offset Correction: v_off_ini = -99.686 uV, v_off_fin = -91.720 uV 12:07:19 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 25.486 uVs 12:07:19 Integrated Voltage Measurement... 12:07:28 Voltage Offset Correction: v_off_ini = -92.977 uV, v_off_fin = -76.997 uV 12:07:28 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -32.728 uVs 12:07:28 Integrated Voltage Measurement... 12:07:37 Voltage Offset Correction: v_off_ini = -99.119 uV, v_off_fin = -108.250 uV 12:07:37 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 40.330 uVs 12:07:39 I2X Average and RMS of Measurements VT = -34.206 +- 8.269 uVs I2X = 3.147 +- 0.761 uTm^2 12:07:42 Wire Position: X0 = 0.0000 mm, Y0 = 0.5000 mm 12:07:44 Integrated Voltage Measurement... 12:07:53 Voltage Offset Correction: v_off_ini = -86.217 uV, v_off_fin = -81.474 uV 12:07:53 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -176.442 uVs 12:07:53 Integrated Voltage Measurement... 12:08:02 Voltage Offset Correction: v_off_ini = -103.938 uV, v_off_fin = -100.518 uV 12:08:02 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 164.481 uVs 12:08:02 Integrated Voltage Measurement... 12:08:11 Voltage Offset Correction: v_off_ini = -95.704 uV, v_off_fin = -90.804 uV 12:08:11 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -186.567 uVs 12:08:11 Integrated Voltage Measurement... 12:08:20 Voltage Offset Correction: v_off_ini = -86.253 uV, v_off_fin = -96.124 uV 12:08:20 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 173.007 uVs 12:08:21 Integrated Voltage Measurement... 12:08:29 Voltage Offset Correction: v_off_ini = -91.266 uV, v_off_fin = -100.823 uV 12:08:30 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -178.330 uVs 12:08:30 Integrated Voltage Measurement... 12:08:39 Voltage Offset Correction: v_off_ini = -108.858 uV, v_off_fin = -114.452 uV 12:08:39 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 193.497 uVs 12:08:41 I2Y Average and RMS of Measurements VT = -178.721 +- 6.353 uVs I2Y = -16.442 +- 0.584 uTm^2 12:08:45 Wire Position: X0 = 0.0000 mm, Y0 = 1.0000 mm 12:08:49 Integrated Voltage Measurement... 12:08:58 Voltage Offset Correction: v_off_ini = -128.497 uV, v_off_fin = -109.815 uV 12:08:58 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -27.903 uVs 12:08:58 Integrated Voltage Measurement... 12:09:07 Voltage Offset Correction: v_off_ini = -93.352 uV, v_off_fin = -97.537 uV 12:09:07 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 16.328 uVs 12:09:07 Integrated Voltage Measurement... 12:09:16 Voltage Offset Correction: v_off_ini = -106.752 uV, v_off_fin = -119.893 uV 12:09:16 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -36.129 uVs 12:09:16 Integrated Voltage Measurement... 12:09:25 Voltage Offset Correction: v_off_ini = -115.811 uV, v_off_fin = -102.438 uV 12:09:25 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 38.701 uVs 12:09:25 Integrated Voltage Measurement... 12:09:34 Voltage Offset Correction: v_off_ini = -103.361 uV, v_off_fin = -117.768 uV 12:09:34 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -17.429 uVs 12:09:34 Integrated Voltage Measurement... 12:09:43 Voltage Offset Correction: v_off_ini = -99.422 uV, v_off_fin = -95.139 uV 12:09:43 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 20.531 uVs 12:09:48 I1X Average and RMS of Measurements VT = -26.170 +- 8.054 uVs I1X = 5.234 +- 1.611 uTm 12:09:50 Wire Position: X0 = 0.0000 mm, Y0 = 1.0000 mm 12:09:54 Integrated Voltage Measurement... 12:10:03 Voltage Offset Correction: v_off_ini = -113.015 uV, v_off_fin = -118.549 uV 12:10:03 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4998 mm dXY = 0.4999 mm VT = -158.389 uVs 12:10:03 Integrated Voltage Measurement... 12:10:12 Voltage Offset Correction: v_off_ini = -101.543 uV, v_off_fin = -111.624 uV 12:10:12 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 172.282 uVs 12:10:12 Integrated Voltage Measurement... 12:10:21 Voltage Offset Correction: v_off_ini = -121.838 uV, v_off_fin = -119.519 uV 12:10:21 Integrated Voltage Measurement dXYus = 0.5001 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -158.314 uVs 12:10:21 Integrated Voltage Measurement... 12:10:30 Voltage Offset Correction: v_off_ini = -117.711 uV, v_off_fin = -113.109 uV 12:10:30 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 166.931 uVs 12:10:30 Integrated Voltage Measurement... 12:10:39 Voltage Offset Correction: v_off_ini = -117.225 uV, v_off_fin = -124.647 uV 12:10:39 Integrated Voltage Measurement dXYus = 0.5001 mm, dXYds = 0.5001 mm dXY = 0.5001 mm VT = -163.244 uVs 12:10:39 Integrated Voltage Measurement... 12:10:48 Voltage Offset Correction: v_off_ini = -118.168 uV, v_off_fin = -108.778 uV 12:10:48 Integrated Voltage Measurement dXYus = -0.5001 mm, dXYds = -0.5001 mm dXY = -0.5001 mm VT = 168.384 uVs 12:10:53 I1Y Average and RMS of Measurements VT = -164.591 +- 1.405 uVs I1Y = -32.916 +- 0.280 uTm 12:10:55 Wire Position: X0 = 0.0000 mm, Y0 = 1.0000 mm 12:10:58 Integrated Voltage Measurement... 12:11:07 Voltage Offset Correction: v_off_ini = -128.754 uV, v_off_fin = -100.165 uV 12:11:07 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -22.082 uVs 12:11:07 Integrated Voltage Measurement... 12:11:16 Voltage Offset Correction: v_off_ini = -110.125 uV, v_off_fin = -110.215 uV 12:11:16 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 23.241 uVs 12:11:16 Integrated Voltage Measurement... 12:11:25 Voltage Offset Correction: v_off_ini = -101.448 uV, v_off_fin = -86.224 uV 12:11:25 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -52.550 uVs 12:11:25 Integrated Voltage Measurement... 12:11:34 Voltage Offset Correction: v_off_ini = -102.341 uV, v_off_fin = -110.569 uV 12:11:34 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 37.227 uVs 12:11:34 Integrated Voltage Measurement... 12:11:43 Voltage Offset Correction: v_off_ini = -112.660 uV, v_off_fin = -107.691 uV 12:11:43 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -39.980 uVs 12:11:43 Integrated Voltage Measurement... 12:11:52 Voltage Offset Correction: v_off_ini = -118.671 uV, v_off_fin = -115.519 uV 12:11:52 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 39.704 uVs 12:11:54 I2X Average and RMS of Measurements VT = -35.797 +- 9.514 uVs I2X = 3.293 +- 0.875 uTm^2 12:11:57 Wire Position: X0 = 0.0000 mm, Y0 = 1.0000 mm 12:11:59 Integrated Voltage Measurement... 12:12:08 Voltage Offset Correction: v_off_ini = -106.044 uV, v_off_fin = -108.089 uV 12:12:08 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -175.054 uVs 12:12:08 Integrated Voltage Measurement... 12:12:17 Voltage Offset Correction: v_off_ini = -108.065 uV, v_off_fin = -92.213 uV 12:12:17 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 183.393 uVs 12:12:18 Integrated Voltage Measurement... 12:12:27 Voltage Offset Correction: v_off_ini = -103.958 uV, v_off_fin = -98.265 uV 12:12:27 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -190.049 uVs 12:12:27 Integrated Voltage Measurement... 12:12:36 Voltage Offset Correction: v_off_ini = -99.780 uV, v_off_fin = -98.199 uV 12:12:36 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 186.773 uVs 12:12:36 Integrated Voltage Measurement... 12:12:45 Voltage Offset Correction: v_off_ini = -97.625 uV, v_off_fin = -98.851 uV 12:12:45 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -176.205 uVs 12:12:45 Integrated Voltage Measurement... 12:12:54 Voltage Offset Correction: v_off_ini = -96.389 uV, v_off_fin = -87.821 uV 12:12:54 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 173.550 uVs 12:12:56 I2Y Average and RMS of Measurements VT = -180.837 +- 5.642 uVs I2Y = -16.637 +- 0.519 uTm^2 12:13:12 Wire Position: X0 = -2.0000 mm, Y0 = 0.0000 mm 12:13:16 Integrated Voltage Measurement... 12:13:25 Voltage Offset Correction: v_off_ini = -86.188 uV, v_off_fin = -95.043 uV 12:13:25 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -30.200 uVs 12:13:25 Integrated Voltage Measurement... 12:13:34 Voltage Offset Correction: v_off_ini = -76.342 uV, v_off_fin = -75.688 uV 12:13:34 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 9.491 uVs 12:13:34 Integrated Voltage Measurement... 12:13:43 Voltage Offset Correction: v_off_ini = -78.653 uV, v_off_fin = -64.084 uV 12:13:43 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -21.178 uVs 12:13:43 Integrated Voltage Measurement... 12:13:52 Voltage Offset Correction: v_off_ini = -73.214 uV, v_off_fin = -74.430 uV 12:13:52 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 6.833 uVs 12:13:52 Integrated Voltage Measurement... 12:14:01 Voltage Offset Correction: v_off_ini = -68.513 uV, v_off_fin = -64.686 uV 12:14:01 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -41.393 uVs 12:14:01 Integrated Voltage Measurement... 12:14:10 Voltage Offset Correction: v_off_ini = -70.344 uV, v_off_fin = -77.493 uV 12:14:10 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 15.766 uVs 12:14:15 I1X Average and RMS of Measurements VT = -20.810 +- 5.989 uVs I1X = 4.162 +- 1.198 uTm 12:14:17 Wire Position: X0 = -2.0000 mm, Y0 = 0.0000 mm 12:14:21 Integrated Voltage Measurement... 12:14:30 Voltage Offset Correction: v_off_ini = -62.124 uV, v_off_fin = -74.064 uV 12:14:30 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -161.471 uVs 12:14:30 Integrated Voltage Measurement... 12:14:39 Voltage Offset Correction: v_off_ini = -67.363 uV, v_off_fin = -61.752 uV 12:14:39 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 161.258 uVs 12:14:39 Integrated Voltage Measurement... 12:14:48 Voltage Offset Correction: v_off_ini = -79.375 uV, v_off_fin = -88.037 uV 12:14:48 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -143.275 uVs 12:14:48 Integrated Voltage Measurement... 12:14:57 Voltage Offset Correction: v_off_ini = -73.533 uV, v_off_fin = -61.250 uV 12:14:57 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.4999 mm dXY = -0.4999 mm VT = 155.742 uVs 12:14:57 Integrated Voltage Measurement... 12:15:06 Voltage Offset Correction: v_off_ini = -59.993 uV, v_off_fin = -78.078 uV 12:15:06 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -140.921 uVs 12:15:06 Integrated Voltage Measurement... 12:15:15 Voltage Offset Correction: v_off_ini = -62.821 uV, v_off_fin = -63.015 uV 12:15:15 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 136.543 uVs 12:15:20 I1Y Average and RMS of Measurements VT = -149.868 +- 9.243 uVs I1Y = -29.975 +- 1.849 uTm 12:15:25 Wire Position: X0 = -2.0000 mm, Y0 = 0.0000 mm 12:15:27 Integrated Voltage Measurement... 12:15:36 Voltage Offset Correction: v_off_ini = -61.200 uV, v_off_fin = -59.357 uV 12:15:36 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -9.463 uVs 12:15:36 Integrated Voltage Measurement... 12:15:45 Voltage Offset Correction: v_off_ini = -74.354 uV, v_off_fin = -74.691 uV 12:15:45 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 17.800 uVs 12:15:45 Integrated Voltage Measurement... 12:15:54 Voltage Offset Correction: v_off_ini = -65.393 uV, v_off_fin = -65.572 uV 12:15:54 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -5.383 uVs 12:15:54 Integrated Voltage Measurement... 12:16:03 Voltage Offset Correction: v_off_ini = -85.665 uV, v_off_fin = -61.683 uV 12:16:03 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 23.624 uVs 12:16:03 Integrated Voltage Measurement... 12:16:12 Voltage Offset Correction: v_off_ini = -65.697 uV, v_off_fin = -64.031 uV 12:16:13 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -26.739 uVs 12:16:13 Integrated Voltage Measurement... 12:16:21 Voltage Offset Correction: v_off_ini = -66.598 uV, v_off_fin = -67.749 uV 12:16:22 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 24.450 uVs 12:16:24 I2X Average and RMS of Measurements VT = -17.910 +- 5.446 uVs I2X = 1.648 +- 0.501 uTm^2 12:16:29 Wire Position: X0 = -2.0000 mm, Y0 = 0.0000 mm 12:16:32 Integrated Voltage Measurement... 12:16:41 Voltage Offset Correction: v_off_ini = -56.730 uV, v_off_fin = -56.749 uV 12:16:41 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -166.387 uVs 12:16:41 Integrated Voltage Measurement... 12:16:50 Voltage Offset Correction: v_off_ini = -55.061 uV, v_off_fin = -54.597 uV 12:16:50 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 181.500 uVs 12:16:50 Integrated Voltage Measurement... 12:16:59 Voltage Offset Correction: v_off_ini = -55.295 uV, v_off_fin = -68.755 uV 12:16:59 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -181.776 uVs 12:16:59 Integrated Voltage Measurement... 12:17:08 Voltage Offset Correction: v_off_ini = -66.599 uV, v_off_fin = -62.479 uV 12:17:08 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 178.368 uVs 12:17:08 Integrated Voltage Measurement... 12:17:17 Voltage Offset Correction: v_off_ini = -58.201 uV, v_off_fin = -62.098 uV 12:17:17 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -160.909 uVs 12:17:17 Integrated Voltage Measurement... 12:17:26 Voltage Offset Correction: v_off_ini = -51.763 uV, v_off_fin = -48.572 uV 12:17:26 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 160.293 uVs 12:17:28 I2Y Average and RMS of Measurements VT = -171.539 +- 8.129 uVs I2Y = -15.782 +- 0.748 uTm^2 12:17:33 Wire Position: X0 = -1.0000 mm, Y0 = 0.0000 mm 12:17:37 Integrated Voltage Measurement... 12:17:46 Voltage Offset Correction: v_off_ini = -54.308 uV, v_off_fin = -43.061 uV 12:17:46 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -30.605 uVs 12:17:46 Integrated Voltage Measurement... 12:17:55 Voltage Offset Correction: v_off_ini = -52.803 uV, v_off_fin = -42.065 uV 12:17:55 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 9.640 uVs 12:17:55 Integrated Voltage Measurement... 12:18:04 Voltage Offset Correction: v_off_ini = -52.041 uV, v_off_fin = -56.230 uV 12:18:04 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -5.725 uVs 12:18:04 Integrated Voltage Measurement... 12:18:13 Voltage Offset Correction: v_off_ini = -47.224 uV, v_off_fin = -56.485 uV 12:18:13 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 14.728 uVs 12:18:13 Integrated Voltage Measurement... 12:18:22 Voltage Offset Correction: v_off_ini = -48.137 uV, v_off_fin = -47.250 uV 12:18:22 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -31.410 uVs 12:18:22 Integrated Voltage Measurement... 12:18:31 Voltage Offset Correction: v_off_ini = -44.713 uV, v_off_fin = -38.685 uV 12:18:31 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 14.496 uVs 12:18:36 I1X Average and RMS of Measurements VT = -17.767 +- 5.456 uVs I1X = 3.554 +- 1.091 uTm 12:18:38 Wire Position: X0 = -1.0000 mm, Y0 = 0.0000 mm 12:18:42 Integrated Voltage Measurement... 12:18:51 Voltage Offset Correction: v_off_ini = -45.120 uV, v_off_fin = -47.368 uV 12:18:51 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -140.066 uVs 12:18:51 Integrated Voltage Measurement... 12:19:00 Voltage Offset Correction: v_off_ini = -55.629 uV, v_off_fin = -31.326 uV 12:19:00 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 173.126 uVs 12:19:00 Integrated Voltage Measurement... 12:19:09 Voltage Offset Correction: v_off_ini = -46.265 uV, v_off_fin = -40.203 uV 12:19:09 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -146.151 uVs 12:19:09 Integrated Voltage Measurement... 12:19:18 Voltage Offset Correction: v_off_ini = -48.721 uV, v_off_fin = -34.153 uV 12:19:18 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 149.847 uVs 12:19:18 Integrated Voltage Measurement... 12:19:27 Voltage Offset Correction: v_off_ini = -49.354 uV, v_off_fin = -49.305 uV 12:19:27 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -145.506 uVs 12:19:27 Integrated Voltage Measurement... 12:19:36 Voltage Offset Correction: v_off_ini = -40.994 uV, v_off_fin = -52.510 uV 12:19:36 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 158.061 uVs 12:19:41 I1Y Average and RMS of Measurements VT = -152.126 +- 3.518 uVs I1Y = -30.425 +- 0.704 uTm 12:19:46 Wire Position: X0 = -1.0000 mm, Y0 = 0.0000 mm 12:19:48 Integrated Voltage Measurement... 12:19:57 Voltage Offset Correction: v_off_ini = -41.546 uV, v_off_fin = -48.318 uV 12:19:57 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -15.769 uVs 12:19:57 Integrated Voltage Measurement... 12:20:06 Voltage Offset Correction: v_off_ini = -54.993 uV, v_off_fin = -52.625 uV 12:20:06 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 17.518 uVs 12:20:06 Integrated Voltage Measurement... 12:20:15 Voltage Offset Correction: v_off_ini = -54.070 uV, v_off_fin = -40.322 uV 12:20:15 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -44.853 uVs 12:20:15 Integrated Voltage Measurement... 12:20:24 Voltage Offset Correction: v_off_ini = -38.004 uV, v_off_fin = -47.001 uV 12:20:24 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 48.743 uVs 12:20:24 Integrated Voltage Measurement... 12:20:33 Voltage Offset Correction: v_off_ini = -40.340 uV, v_off_fin = -40.062 uV 12:20:33 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -53.941 uVs 12:20:33 Integrated Voltage Measurement... 12:20:42 Voltage Offset Correction: v_off_ini = -39.388 uV, v_off_fin = -44.136 uV 12:20:42 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 27.035 uVs 12:20:45 I2X Average and RMS of Measurements VT = -34.643 +- 12.986 uVs I2X = 3.187 +- 1.195 uTm^2 12:20:50 Wire Position: X0 = -1.0000 mm, Y0 = 0.0000 mm 12:20:53 Integrated Voltage Measurement... 12:21:02 Voltage Offset Correction: v_off_ini = -42.083 uV, v_off_fin = -41.183 uV 12:21:02 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -148.301 uVs 12:21:02 Integrated Voltage Measurement... 12:21:11 Voltage Offset Correction: v_off_ini = -54.075 uV, v_off_fin = -39.858 uV 12:21:11 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 157.523 uVs 12:21:11 Integrated Voltage Measurement... 12:21:20 Voltage Offset Correction: v_off_ini = -41.605 uV, v_off_fin = -54.585 uV 12:21:20 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = -162.933 uVs 12:21:20 Integrated Voltage Measurement... 12:21:29 Voltage Offset Correction: v_off_ini = -43.272 uV, v_off_fin = -30.383 uV 12:21:29 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 168.378 uVs 12:21:29 Integrated Voltage Measurement... 12:21:38 Voltage Offset Correction: v_off_ini = -47.350 uV, v_off_fin = -36.802 uV 12:21:38 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -160.694 uVs 12:21:38 Integrated Voltage Measurement... 12:21:47 Voltage Offset Correction: v_off_ini = -55.083 uV, v_off_fin = -39.240 uV 12:21:47 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 169.169 uVs 12:21:49 I2Y Average and RMS of Measurements VT = -161.166 +- 5.844 uVs I2Y = -14.828 +- 0.538 uTm^2 12:21:54 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 12:21:58 Integrated Voltage Measurement... 12:22:07 Voltage Offset Correction: v_off_ini = -32.150 uV, v_off_fin = -39.897 uV 12:22:07 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -37.569 uVs 12:22:07 Integrated Voltage Measurement... 12:22:16 Voltage Offset Correction: v_off_ini = -34.712 uV, v_off_fin = -24.510 uV 12:22:16 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 13.002 uVs 12:22:16 Integrated Voltage Measurement... 12:22:25 Voltage Offset Correction: v_off_ini = -28.419 uV, v_off_fin = -41.153 uV 12:22:25 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -34.380 uVs 12:22:25 Integrated Voltage Measurement... 12:22:34 Voltage Offset Correction: v_off_ini = -18.313 uV, v_off_fin = -21.635 uV 12:22:34 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 19.792 uVs 12:22:34 Integrated Voltage Measurement... 12:22:43 Voltage Offset Correction: v_off_ini = -29.899 uV, v_off_fin = -27.072 uV 12:22:43 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -11.645 uVs 12:22:43 Integrated Voltage Measurement... 12:22:52 Voltage Offset Correction: v_off_ini = -30.068 uV, v_off_fin = -38.298 uV 12:22:52 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 39.896 uVs 12:22:56 I1X Average and RMS of Measurements VT = -26.047 +- 0.761 uVs I1X = 5.210 +- 0.152 uTm 12:22:59 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 12:23:03 Integrated Voltage Measurement... 12:23:12 Voltage Offset Correction: v_off_ini = -14.429 uV, v_off_fin = -29.660 uV 12:23:12 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -156.168 uVs 12:23:12 Integrated Voltage Measurement... 12:23:21 Voltage Offset Correction: v_off_ini = -13.841 uV, v_off_fin = -17.532 uV 12:23:21 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.4999 mm dXY = -0.4999 mm VT = 144.656 uVs 12:23:21 Integrated Voltage Measurement... 12:23:30 Voltage Offset Correction: v_off_ini = -28.638 uV, v_off_fin = -25.519 uV 12:23:30 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.4999 mm dXY = 0.4999 mm VT = -135.694 uVs 12:23:30 Integrated Voltage Measurement... 12:23:39 Voltage Offset Correction: v_off_ini = -22.411 uV, v_off_fin = -3.785 uV 12:23:39 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 131.616 uVs 12:23:39 Integrated Voltage Measurement... 12:23:48 Voltage Offset Correction: v_off_ini = -7.971 uV, v_off_fin = -25.897 uV 12:23:48 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -151.124 uVs 12:23:48 Integrated Voltage Measurement... 12:23:57 Voltage Offset Correction: v_off_ini = -20.660 uV, v_off_fin = -7.144 uV 12:23:57 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 156.753 uVs 12:24:01 I1Y Average and RMS of Measurements VT = -146.002 +- 8.848 uVs I1Y = -29.202 +- 1.769 uTm 12:24:07 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 12:24:09 Integrated Voltage Measurement... 12:24:18 Voltage Offset Correction: v_off_ini = -12.514 uV, v_off_fin = -10.716 uV 12:24:18 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -52.557 uVs 12:24:18 Integrated Voltage Measurement... 12:24:27 Voltage Offset Correction: v_off_ini = -15.628 uV, v_off_fin = -35.623 uV 12:24:27 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 37.241 uVs 12:24:27 Integrated Voltage Measurement... 12:24:36 Voltage Offset Correction: v_off_ini = -24.678 uV, v_off_fin = -24.534 uV 12:24:36 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -53.030 uVs 12:24:36 Integrated Voltage Measurement... 12:24:45 Voltage Offset Correction: v_off_ini = -28.149 uV, v_off_fin = -22.109 uV 12:24:45 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 59.977 uVs 12:24:45 Integrated Voltage Measurement... 12:24:54 Voltage Offset Correction: v_off_ini = -13.340 uV, v_off_fin = -5.392 uV 12:24:54 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -58.085 uVs 12:24:54 Integrated Voltage Measurement... 12:25:03 Voltage Offset Correction: v_off_ini = -11.454 uV, v_off_fin = -25.198 uV 12:25:03 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 51.294 uVs 12:25:06 I2X Average and RMS of Measurements VT = -52.030 +- 5.097 uVs I2X = 4.787 +- 0.469 uTm^2 12:25:11 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 12:25:13 Integrated Voltage Measurement... 12:25:22 Voltage Offset Correction: v_off_ini = -15.636 uV, v_off_fin = -17.042 uV 12:25:23 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -161.170 uVs 12:25:23 Integrated Voltage Measurement... 12:25:32 Voltage Offset Correction: v_off_ini = -9.568 uV, v_off_fin = -24.984 uV 12:25:32 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 164.479 uVs 12:25:32 Integrated Voltage Measurement... 12:25:41 Voltage Offset Correction: v_off_ini = -28.504 uV, v_off_fin = -28.346 uV 12:25:41 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -161.501 uVs 12:25:41 Integrated Voltage Measurement... 12:25:50 Voltage Offset Correction: v_off_ini = -34.137 uV, v_off_fin = -22.862 uV 12:25:50 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 185.754 uVs 12:25:50 Integrated Voltage Measurement... 12:25:59 Voltage Offset Correction: v_off_ini = -31.085 uV, v_off_fin = -29.930 uV 12:25:59 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -167.433 uVs 12:25:59 Integrated Voltage Measurement... 12:26:08 Voltage Offset Correction: v_off_ini = -33.582 uV, v_off_fin = -24.385 uV 12:26:08 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 169.343 uVs 12:26:10 I2Y Average and RMS of Measurements VT = -168.280 +- 4.411 uVs I2Y = -15.482 +- 0.406 uTm^2 12:26:15 Wire Position: X0 = 1.0000 mm, Y0 = 0.0000 mm 12:26:19 Integrated Voltage Measurement... 12:26:28 Voltage Offset Correction: v_off_ini = -26.830 uV, v_off_fin = -14.899 uV 12:26:28 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -41.639 uVs 12:26:28 Integrated Voltage Measurement... 12:26:37 Voltage Offset Correction: v_off_ini = -21.720 uV, v_off_fin = -16.116 uV 12:26:37 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 31.228 uVs 12:26:37 Integrated Voltage Measurement... 12:26:46 Voltage Offset Correction: v_off_ini = -15.401 uV, v_off_fin = -40.679 uV 12:26:46 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -52.764 uVs 12:26:46 Integrated Voltage Measurement... 12:26:55 Voltage Offset Correction: v_off_ini = -21.068 uV, v_off_fin = -33.324 uV 12:26:55 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 39.619 uVs 12:26:55 Integrated Voltage Measurement... 12:27:04 Voltage Offset Correction: v_off_ini = -29.038 uV, v_off_fin = -34.507 uV 12:27:04 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -27.839 uVs 12:27:04 Integrated Voltage Measurement... 12:27:13 Voltage Offset Correction: v_off_ini = -27.403 uV, v_off_fin = -32.114 uV 12:27:13 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 21.971 uVs 12:27:17 I1X Average and RMS of Measurements VT = -35.843 +- 8.700 uVs I1X = 7.169 +- 1.740 uTm 12:27:20 Wire Position: X0 = 1.0000 mm, Y0 = 0.0000 mm 12:27:24 Integrated Voltage Measurement... 12:27:33 Voltage Offset Correction: v_off_ini = -38.676 uV, v_off_fin = -57.034 uV 12:27:33 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -152.955 uVs 12:27:33 Integrated Voltage Measurement... 12:27:42 Voltage Offset Correction: v_off_ini = -38.558 uV, v_off_fin = -41.020 uV 12:27:42 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 150.239 uVs 12:27:42 Integrated Voltage Measurement... 12:27:51 Voltage Offset Correction: v_off_ini = -44.390 uV, v_off_fin = -45.669 uV 12:27:51 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -154.689 uVs 12:27:51 Integrated Voltage Measurement... 12:28:00 Voltage Offset Correction: v_off_ini = -65.934 uV, v_off_fin = -55.375 uV 12:28:00 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 147.954 uVs 12:28:00 Integrated Voltage Measurement... 12:28:09 Voltage Offset Correction: v_off_ini = -50.622 uV, v_off_fin = -46.429 uV 12:28:09 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -162.804 uVs 12:28:09 Integrated Voltage Measurement... 12:28:18 Voltage Offset Correction: v_off_ini = -47.415 uV, v_off_fin = -29.421 uV 12:28:18 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.4999 mm dXY = -0.4999 mm VT = 162.231 uVs 12:28:22 I1Y Average and RMS of Measurements VT = -155.145 +- 5.214 uVs I1Y = -31.030 +- 1.044 uTm 12:28:27 Wire Position: X0 = 1.0000 mm, Y0 = 0.0000 mm 12:28:30 Integrated Voltage Measurement... 12:28:39 Voltage Offset Correction: v_off_ini = -50.028 uV, v_off_fin = -51.920 uV 12:28:39 Integrated Voltage Measurement dXYus = -1.0001 mm, dXYds = 0.0000 mm dXY = -1.0001 mm VT = -81.239 uVs 12:28:39 Integrated Voltage Measurement... 12:28:48 Voltage Offset Correction: v_off_ini = -51.775 uV, v_off_fin = -68.494 uV 12:28:48 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 62.686 uVs 12:28:48 Integrated Voltage Measurement... 12:28:57 Voltage Offset Correction: v_off_ini = -53.590 uV, v_off_fin = -52.149 uV 12:28:57 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -61.653 uVs 12:28:57 Integrated Voltage Measurement... 12:29:06 Voltage Offset Correction: v_off_ini = -56.725 uV, v_off_fin = -51.042 uV 12:29:06 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 59.069 uVs 12:29:06 Integrated Voltage Measurement... 12:29:15 Voltage Offset Correction: v_off_ini = -54.110 uV, v_off_fin = -45.437 uV 12:29:15 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -37.426 uVs 12:29:15 Integrated Voltage Measurement... 12:29:24 Voltage Offset Correction: v_off_ini = -62.130 uV, v_off_fin = -54.943 uV 12:29:24 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 41.813 uVs 12:29:27 I2X Average and RMS of Measurements VT = -57.314 +- 13.378 uVs I2X = 5.273 +- 1.231 uTm^2 12:29:32 Wire Position: X0 = 1.0000 mm, Y0 = 0.0000 mm 12:29:34 Integrated Voltage Measurement... 12:29:43 Voltage Offset Correction: v_off_ini = -68.262 uV, v_off_fin = -66.361 uV 12:29:43 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -166.541 uVs 12:29:44 Integrated Voltage Measurement... 12:29:52 Voltage Offset Correction: v_off_ini = -57.710 uV, v_off_fin = -63.866 uV 12:29:53 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 163.993 uVs 12:29:53 Integrated Voltage Measurement... 12:30:02 Voltage Offset Correction: v_off_ini = -59.950 uV, v_off_fin = -57.559 uV 12:30:02 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -165.832 uVs 12:30:02 Integrated Voltage Measurement... 12:30:11 Voltage Offset Correction: v_off_ini = -59.323 uV, v_off_fin = -48.037 uV 12:30:11 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 178.688 uVs 12:30:11 Integrated Voltage Measurement... 12:30:20 Voltage Offset Correction: v_off_ini = -48.519 uV, v_off_fin = -54.906 uV 12:30:20 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -174.558 uVs 12:30:20 Integrated Voltage Measurement... 12:30:29 Voltage Offset Correction: v_off_ini = -39.474 uV, v_off_fin = -54.478 uV 12:30:29 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 163.260 uVs 12:30:31 I2Y Average and RMS of Measurements VT = -168.812 +- 2.856 uVs I2Y = -15.531 +- 0.263 uTm^2 12:30:36 Wire Position: X0 = 2.0000 mm, Y0 = 0.0000 mm 12:30:40 Integrated Voltage Measurement... 12:30:49 Voltage Offset Correction: v_off_ini = -53.757 uV, v_off_fin = -43.381 uV 12:30:49 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -27.945 uVs 12:30:49 Integrated Voltage Measurement... 12:30:58 Voltage Offset Correction: v_off_ini = -59.783 uV, v_off_fin = -43.900 uV 12:30:58 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 57.738 uVs 12:30:58 Integrated Voltage Measurement... 12:31:07 Voltage Offset Correction: v_off_ini = -46.825 uV, v_off_fin = -52.966 uV 12:31:07 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -36.432 uVs 12:31:07 Integrated Voltage Measurement... 12:31:16 Voltage Offset Correction: v_off_ini = -68.216 uV, v_off_fin = -53.414 uV 12:31:16 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 40.743 uVs 12:31:16 Integrated Voltage Measurement... 12:31:25 Voltage Offset Correction: v_off_ini = -60.776 uV, v_off_fin = -61.014 uV 12:31:25 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -49.452 uVs 12:31:25 Integrated Voltage Measurement... 12:31:34 Voltage Offset Correction: v_off_ini = -63.735 uV, v_off_fin = -60.039 uV 12:31:34 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 30.621 uVs 12:31:38 I1X Average and RMS of Measurements VT = -40.488 +- 1.766 uVs I1X = 8.098 +- 0.353 uTm 12:31:41 Wire Position: X0 = 2.0000 mm, Y0 = 0.0000 mm 12:31:45 Integrated Voltage Measurement... 12:31:54 Voltage Offset Correction: v_off_ini = -56.064 uV, v_off_fin = -74.037 uV 12:31:54 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.4999 mm dXY = 0.4999 mm VT = -139.756 uVs 12:31:54 Integrated Voltage Measurement... 12:32:03 Voltage Offset Correction: v_off_ini = -64.569 uV, v_off_fin = -68.030 uV 12:32:03 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 159.399 uVs 12:32:03 Integrated Voltage Measurement... 12:32:12 Voltage Offset Correction: v_off_ini = -67.907 uV, v_off_fin = -67.349 uV 12:32:12 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -138.624 uVs 12:32:12 Integrated Voltage Measurement... 12:32:21 Voltage Offset Correction: v_off_ini = -68.246 uV, v_off_fin = -51.524 uV 12:32:21 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 150.312 uVs 12:32:21 Integrated Voltage Measurement... 12:32:30 Voltage Offset Correction: v_off_ini = -52.199 uV, v_off_fin = -68.763 uV 12:32:30 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -159.043 uVs 12:32:30 Integrated Voltage Measurement... 12:32:39 Voltage Offset Correction: v_off_ini = -61.547 uV, v_off_fin = -51.812 uV 12:32:39 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 138.896 uVs 12:32:43 I1Y Average and RMS of Measurements VT = -147.671 +- 2.279 uVs I1Y = -29.535 +- 0.457 uTm 12:32:48 Wire Position: X0 = 2.0000 mm, Y0 = 0.0000 mm 12:32:51 Integrated Voltage Measurement... 12:33:00 Voltage Offset Correction: v_off_ini = -51.029 uV, v_off_fin = -53.602 uV 12:33:00 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -67.033 uVs 12:33:00 Integrated Voltage Measurement... 12:33:09 Voltage Offset Correction: v_off_ini = -58.526 uV, v_off_fin = -64.455 uV 12:33:09 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 76.701 uVs 12:33:09 Integrated Voltage Measurement... 12:33:18 Voltage Offset Correction: v_off_ini = -67.293 uV, v_off_fin = -68.997 uV 12:33:18 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -46.594 uVs 12:33:18 Integrated Voltage Measurement... 12:33:27 Voltage Offset Correction: v_off_ini = -73.869 uV, v_off_fin = -76.235 uV 12:33:27 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 63.868 uVs 12:33:27 Integrated Voltage Measurement... 12:33:36 Voltage Offset Correction: v_off_ini = -78.566 uV, v_off_fin = -68.730 uV 12:33:36 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -45.851 uVs 12:33:36 Integrated Voltage Measurement... 12:33:45 Voltage Offset Correction: v_off_ini = -64.102 uV, v_off_fin = -82.437 uV 12:33:45 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 47.237 uVs 12:33:48 I2X Average and RMS of Measurements VT = -57.881 +- 10.506 uVs I2X = 5.325 +- 0.967 uTm^2 12:33:53 Wire Position: X0 = 2.0000 mm, Y0 = 0.0000 mm 12:33:55 Integrated Voltage Measurement... 12:34:04 Voltage Offset Correction: v_off_ini = -60.363 uV, v_off_fin = -65.978 uV 12:34:04 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -173.903 uVs 12:34:04 Integrated Voltage Measurement... 12:34:13 Voltage Offset Correction: v_off_ini = -70.165 uV, v_off_fin = -80.385 uV 12:34:13 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 176.592 uVs 12:34:13 Integrated Voltage Measurement... 12:34:22 Voltage Offset Correction: v_off_ini = -73.919 uV, v_off_fin = -64.066 uV 12:34:22 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -164.888 uVs 12:34:23 Integrated Voltage Measurement... 12:34:31 Voltage Offset Correction: v_off_ini = -68.300 uV, v_off_fin = -81.133 uV 12:34:32 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 164.173 uVs 12:34:32 Integrated Voltage Measurement... 12:34:41 Voltage Offset Correction: v_off_ini = -72.872 uV, v_off_fin = -85.845 uV 12:34:41 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -171.589 uVs 12:34:41 Integrated Voltage Measurement... 12:34:50 Voltage Offset Correction: v_off_ini = -66.907 uV, v_off_fin = -82.568 uV 12:34:50 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 157.595 uVs 12:34:52 I2Y Average and RMS of Measurements VT = -168.123 +- 5.038 uVs I2Y = -15.467 +- 0.463 uTm^2 12:35:00 Wire Position: X0 = 0.0000 mm, Y0 = -1.0000 mm 12:35:04 Integrated Voltage Measurement... 12:35:13 Voltage Offset Correction: v_off_ini = -85.662 uV, v_off_fin = -80.704 uV 12:35:13 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -23.271 uVs 12:35:13 Integrated Voltage Measurement... 12:35:22 Voltage Offset Correction: v_off_ini = -80.612 uV, v_off_fin = -70.156 uV 12:35:22 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 16.987 uVs 12:35:22 Integrated Voltage Measurement... 12:35:31 Voltage Offset Correction: v_off_ini = -65.063 uV, v_off_fin = -81.331 uV 12:35:31 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -31.912 uVs 12:35:31 Integrated Voltage Measurement... 12:35:40 Voltage Offset Correction: v_off_ini = -64.054 uV, v_off_fin = -65.899 uV 12:35:40 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 22.853 uVs 12:35:40 Integrated Voltage Measurement... 12:35:49 Voltage Offset Correction: v_off_ini = -69.106 uV, v_off_fin = -74.337 uV 12:35:49 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -41.301 uVs 12:35:49 Integrated Voltage Measurement... 12:35:58 Voltage Offset Correction: v_off_ini = -71.130 uV, v_off_fin = -60.535 uV 12:35:58 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 37.485 uVs 12:36:02 I1X Average and RMS of Measurements VT = -28.968 +- 7.944 uVs I1X = 5.794 +- 1.589 uTm 12:36:05 Wire Position: X0 = 0.0000 mm, Y0 = -1.0000 mm 12:36:09 Integrated Voltage Measurement... 12:36:18 Voltage Offset Correction: v_off_ini = -74.945 uV, v_off_fin = -64.654 uV 12:36:18 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -128.754 uVs 12:36:18 Integrated Voltage Measurement... 12:36:27 Voltage Offset Correction: v_off_ini = -56.351 uV, v_off_fin = -44.717 uV 12:36:27 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 144.053 uVs 12:36:27 Integrated Voltage Measurement... 12:36:36 Voltage Offset Correction: v_off_ini = -74.907 uV, v_off_fin = -71.182 uV 12:36:36 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -152.225 uVs 12:36:36 Integrated Voltage Measurement... 12:36:45 Voltage Offset Correction: v_off_ini = -64.636 uV, v_off_fin = -57.770 uV 12:36:45 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 141.573 uVs 12:36:45 Integrated Voltage Measurement... 12:36:54 Voltage Offset Correction: v_off_ini = -53.710 uV, v_off_fin = -60.483 uV 12:36:54 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -160.121 uVs 12:36:54 Integrated Voltage Measurement... 12:37:03 Voltage Offset Correction: v_off_ini = -70.654 uV, v_off_fin = -63.725 uV 12:37:03 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 150.516 uVs 12:37:07 I1Y Average and RMS of Measurements VT = -146.207 +- 7.737 uVs I1Y = -29.241 +- 1.547 uTm 12:37:10 Wire Position: X0 = 0.0000 mm, Y0 = -1.0000 mm 12:37:12 Integrated Voltage Measurement... 12:37:21 Voltage Offset Correction: v_off_ini = -68.737 uV, v_off_fin = -53.590 uV 12:37:22 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -30.914 uVs 12:37:22 Integrated Voltage Measurement... 12:37:31 Voltage Offset Correction: v_off_ini = -50.871 uV, v_off_fin = -56.777 uV 12:37:31 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 35.316 uVs 12:37:31 Integrated Voltage Measurement... 12:37:40 Voltage Offset Correction: v_off_ini = -72.194 uV, v_off_fin = -50.465 uV 12:37:40 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -25.585 uVs 12:37:40 Integrated Voltage Measurement... 12:37:49 Voltage Offset Correction: v_off_ini = -55.141 uV, v_off_fin = -56.131 uV 12:37:49 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 30.963 uVs 12:37:49 Integrated Voltage Measurement... 12:37:58 Voltage Offset Correction: v_off_ini = -50.043 uV, v_off_fin = -56.726 uV 12:37:58 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -44.856 uVs 12:37:58 Integrated Voltage Measurement... 12:38:07 Voltage Offset Correction: v_off_ini = -70.099 uV, v_off_fin = -71.364 uV 12:38:07 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 52.236 uVs 12:38:09 I2X Average and RMS of Measurements VT = -36.645 +- 8.644 uVs I2X = 3.371 +- 0.795 uTm^2 12:38:12 Wire Position: X0 = 0.0000 mm, Y0 = -1.0000 mm 12:38:14 Integrated Voltage Measurement... 12:38:23 Voltage Offset Correction: v_off_ini = -43.092 uV, v_off_fin = -50.000 uV 12:38:23 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -176.024 uVs 12:38:23 Integrated Voltage Measurement... 12:38:32 Voltage Offset Correction: v_off_ini = -45.657 uV, v_off_fin = -47.810 uV 12:38:32 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 168.831 uVs 12:38:32 Integrated Voltage Measurement... 12:38:41 Voltage Offset Correction: v_off_ini = -49.845 uV, v_off_fin = -44.711 uV 12:38:41 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -165.435 uVs 12:38:41 Integrated Voltage Measurement... 12:38:50 Voltage Offset Correction: v_off_ini = -41.670 uV, v_off_fin = -30.240 uV 12:38:50 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 192.449 uVs 12:38:50 Integrated Voltage Measurement... 12:38:59 Voltage Offset Correction: v_off_ini = -30.762 uV, v_off_fin = -42.188 uV 12:38:59 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -159.646 uVs 12:38:59 Integrated Voltage Measurement... 12:39:08 Voltage Offset Correction: v_off_ini = -40.608 uV, v_off_fin = -24.431 uV 12:39:08 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 151.418 uVs 12:39:11 I2Y Average and RMS of Measurements VT = -168.967 +- 9.865 uVs I2Y = -15.545 +- 0.908 uTm^2 12:39:15 Wire Position: X0 = 0.0000 mm, Y0 = -0.5000 mm 12:39:19 Integrated Voltage Measurement... 12:39:28 Voltage Offset Correction: v_off_ini = -31.436 uV, v_off_fin = -44.042 uV 12:39:28 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -32.574 uVs 12:39:28 Integrated Voltage Measurement... 12:39:37 Voltage Offset Correction: v_off_ini = -30.593 uV, v_off_fin = -38.834 uV 12:39:37 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 40.098 uVs 12:39:37 Integrated Voltage Measurement... 12:39:46 Voltage Offset Correction: v_off_ini = -41.045 uV, v_off_fin = -45.520 uV 12:39:46 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -41.062 uVs 12:39:46 Integrated Voltage Measurement... 12:39:55 Voltage Offset Correction: v_off_ini = -27.813 uV, v_off_fin = -52.396 uV 12:39:55 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 36.441 uVs 12:39:55 Integrated Voltage Measurement... 12:40:04 Voltage Offset Correction: v_off_ini = -44.491 uV, v_off_fin = -52.364 uV 12:40:04 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -28.785 uVs 12:40:04 Integrated Voltage Measurement... 12:40:13 Voltage Offset Correction: v_off_ini = -38.513 uV, v_off_fin = -51.839 uV 12:40:13 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 13.544 uVs 12:40:18 I1X Average and RMS of Measurements VT = -32.084 +- 7.784 uVs I1X = 6.417 +- 1.557 uTm 12:40:20 Wire Position: X0 = 0.0000 mm, Y0 = -0.5000 mm 12:40:24 Integrated Voltage Measurement... 12:40:33 Voltage Offset Correction: v_off_ini = -51.126 uV, v_off_fin = -63.885 uV 12:40:33 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -151.637 uVs 12:40:33 Integrated Voltage Measurement... 12:40:42 Voltage Offset Correction: v_off_ini = -48.467 uV, v_off_fin = -50.640 uV 12:40:42 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 170.052 uVs 12:40:42 Integrated Voltage Measurement... 12:40:51 Voltage Offset Correction: v_off_ini = -56.334 uV, v_off_fin = -62.643 uV 12:40:51 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -139.736 uVs 12:40:51 Integrated Voltage Measurement... 12:41:00 Voltage Offset Correction: v_off_ini = -51.022 uV, v_off_fin = -58.761 uV 12:41:00 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 147.026 uVs 12:41:00 Integrated Voltage Measurement... 12:41:09 Voltage Offset Correction: v_off_ini = -71.410 uV, v_off_fin = -71.013 uV 12:41:09 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -146.769 uVs 12:41:09 Integrated Voltage Measurement... 12:41:18 Voltage Offset Correction: v_off_ini = -55.513 uV, v_off_fin = -48.833 uV 12:41:18 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 152.058 uVs 12:41:22 I1Y Average and RMS of Measurements VT = -151.213 +- 7.242 uVs I1Y = -30.243 +- 1.448 uTm 12:41:25 Wire Position: X0 = 0.0000 mm, Y0 = -0.5000 mm 12:41:28 Integrated Voltage Measurement... 12:41:37 Voltage Offset Correction: v_off_ini = -37.031 uV, v_off_fin = -47.182 uV 12:41:37 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -32.235 uVs 12:41:37 Integrated Voltage Measurement... 12:41:46 Voltage Offset Correction: v_off_ini = -57.233 uV, v_off_fin = -68.675 uV 12:41:46 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 23.285 uVs 12:41:46 Integrated Voltage Measurement... 12:41:55 Voltage Offset Correction: v_off_ini = -52.634 uV, v_off_fin = -49.747 uV 12:41:55 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -26.177 uVs 12:41:55 Integrated Voltage Measurement... 12:42:04 Voltage Offset Correction: v_off_ini = -67.177 uV, v_off_fin = -72.740 uV 12:42:04 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 29.802 uVs 12:42:04 Integrated Voltage Measurement... 12:42:13 Voltage Offset Correction: v_off_ini = -63.094 uV, v_off_fin = -68.595 uV 12:42:13 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -28.302 uVs 12:42:13 Integrated Voltage Measurement... 12:42:22 Voltage Offset Correction: v_off_ini = -68.707 uV, v_off_fin = -64.607 uV 12:42:22 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 30.438 uVs 12:42:24 I2X Average and RMS of Measurements VT = -28.373 +- 0.711 uVs I2X = 2.610 +- 0.065 uTm^2 12:42:27 Wire Position: X0 = 0.0000 mm, Y0 = -0.5000 mm 12:42:29 Integrated Voltage Measurement... 12:42:38 Voltage Offset Correction: v_off_ini = -59.673 uV, v_off_fin = -58.647 uV 12:42:38 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -171.675 uVs 12:42:38 Integrated Voltage Measurement... 12:42:47 Voltage Offset Correction: v_off_ini = -47.490 uV, v_off_fin = -60.139 uV 12:42:47 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 156.241 uVs 12:42:47 Integrated Voltage Measurement... 12:42:56 Voltage Offset Correction: v_off_ini = -54.856 uV, v_off_fin = -57.170 uV 12:42:56 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -178.114 uVs 12:42:56 Integrated Voltage Measurement... 12:43:05 Voltage Offset Correction: v_off_ini = -68.092 uV, v_off_fin = -71.945 uV 12:43:05 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 171.629 uVs 12:43:06 Integrated Voltage Measurement... 12:43:14 Voltage Offset Correction: v_off_ini = -70.944 uV, v_off_fin = -66.464 uV 12:43:15 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -186.008 uVs 12:43:15 Integrated Voltage Measurement... 12:43:24 Voltage Offset Correction: v_off_ini = -70.630 uV, v_off_fin = -67.718 uV 12:43:24 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 170.122 uVs 12:43:26 I2Y Average and RMS of Measurements VT = -172.298 +- 6.040 uVs I2Y = -15.851 +- 0.556 uTm^2 12:43:30 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 12:43:34 Integrated Voltage Measurement... 12:43:43 Voltage Offset Correction: v_off_ini = -68.412 uV, v_off_fin = -79.138 uV 12:43:43 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -35.029 uVs 12:43:43 Integrated Voltage Measurement... 12:43:52 Voltage Offset Correction: v_off_ini = -72.990 uV, v_off_fin = -70.792 uV 12:43:52 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 33.951 uVs 12:43:52 Integrated Voltage Measurement... 12:44:01 Voltage Offset Correction: v_off_ini = -70.260 uV, v_off_fin = -73.943 uV 12:44:01 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -21.251 uVs 12:44:01 Integrated Voltage Measurement... 12:44:10 Voltage Offset Correction: v_off_ini = -64.196 uV, v_off_fin = -52.926 uV 12:44:10 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 23.624 uVs 12:44:10 Integrated Voltage Measurement... 12:44:19 Voltage Offset Correction: v_off_ini = -85.557 uV, v_off_fin = -70.767 uV 12:44:19 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -14.908 uVs 12:44:19 Integrated Voltage Measurement... 12:44:28 Voltage Offset Correction: v_off_ini = -73.281 uV, v_off_fin = -58.830 uV 12:44:28 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 35.114 uVs 12:44:33 I1X Average and RMS of Measurements VT = -27.313 +- 5.182 uVs I1X = 5.463 +- 1.037 uTm 12:44:35 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 12:44:39 Integrated Voltage Measurement... 12:44:48 Voltage Offset Correction: v_off_ini = -55.878 uV, v_off_fin = -67.532 uV 12:44:48 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -154.707 uVs 12:44:48 Integrated Voltage Measurement... 12:44:57 Voltage Offset Correction: v_off_ini = -72.145 uV, v_off_fin = -62.784 uV 12:44:57 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 147.192 uVs 12:44:57 Integrated Voltage Measurement... 12:45:06 Voltage Offset Correction: v_off_ini = -73.416 uV, v_off_fin = -78.252 uV 12:45:06 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -135.115 uVs 12:45:06 Integrated Voltage Measurement... 12:45:15 Voltage Offset Correction: v_off_ini = -53.308 uV, v_off_fin = -59.623 uV 12:45:15 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 161.594 uVs 12:45:15 Integrated Voltage Measurement... 12:45:24 Voltage Offset Correction: v_off_ini = -53.501 uV, v_off_fin = -77.622 uV 12:45:24 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -152.737 uVs 12:45:24 Integrated Voltage Measurement... 12:45:33 Voltage Offset Correction: v_off_ini = -61.099 uV, v_off_fin = -52.510 uV 12:45:33 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 145.370 uVs 12:45:38 I1Y Average and RMS of Measurements VT = -149.453 +- 1.096 uVs I1Y = -29.890 +- 0.220 uTm 12:45:40 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 12:45:43 Integrated Voltage Measurement... 12:45:52 Voltage Offset Correction: v_off_ini = -68.491 uV, v_off_fin = -53.356 uV 12:45:52 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -34.834 uVs 12:45:52 Integrated Voltage Measurement... 12:46:01 Voltage Offset Correction: v_off_ini = -59.107 uV, v_off_fin = -74.213 uV 12:46:01 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 28.756 uVs 12:46:01 Integrated Voltage Measurement... 12:46:10 Voltage Offset Correction: v_off_ini = -75.812 uV, v_off_fin = -67.176 uV 12:46:10 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -49.047 uVs 12:46:10 Integrated Voltage Measurement... 12:46:19 Voltage Offset Correction: v_off_ini = -57.714 uV, v_off_fin = -62.126 uV 12:46:19 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 27.314 uVs 12:46:19 Integrated Voltage Measurement... 12:46:28 Voltage Offset Correction: v_off_ini = -74.362 uV, v_off_fin = -50.372 uV 12:46:28 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -32.545 uVs 12:46:28 Integrated Voltage Measurement... 12:46:37 Voltage Offset Correction: v_off_ini = -60.585 uV, v_off_fin = -64.447 uV 12:46:37 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 29.860 uVs 12:46:39 I2X Average and RMS of Measurements VT = -33.726 +- 3.159 uVs I2X = 3.103 +- 0.291 uTm^2 12:46:42 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 12:46:44 Integrated Voltage Measurement... 12:46:53 Voltage Offset Correction: v_off_ini = -47.428 uV, v_off_fin = -47.354 uV 12:46:53 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -185.397 uVs 12:46:53 Integrated Voltage Measurement... 12:47:02 Voltage Offset Correction: v_off_ini = -55.850 uV, v_off_fin = -57.216 uV 12:47:02 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 161.104 uVs 12:47:02 Integrated Voltage Measurement... 12:47:11 Voltage Offset Correction: v_off_ini = -45.539 uV, v_off_fin = -45.960 uV 12:47:11 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -181.876 uVs 12:47:11 Integrated Voltage Measurement... 12:47:20 Voltage Offset Correction: v_off_ini = -46.195 uV, v_off_fin = -49.058 uV 12:47:20 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 169.991 uVs 12:47:21 Integrated Voltage Measurement... 12:47:29 Voltage Offset Correction: v_off_ini = -49.983 uV, v_off_fin = -70.985 uV 12:47:30 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -174.349 uVs 12:47:30 Integrated Voltage Measurement... 12:47:39 Voltage Offset Correction: v_off_ini = -64.223 uV, v_off_fin = -64.110 uV 12:47:39 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 182.153 uVs 12:47:41 I2Y Average and RMS of Measurements VT = -175.812 +- 2.043 uVs I2Y = -16.175 +- 0.188 uTm^2 12:47:45 Wire Position: X0 = 0.0000 mm, Y0 = 0.5000 mm 12:47:49 Integrated Voltage Measurement... 12:47:58 Voltage Offset Correction: v_off_ini = -75.218 uV, v_off_fin = -71.251 uV 12:47:58 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -38.547 uVs 12:47:58 Integrated Voltage Measurement... 12:48:07 Voltage Offset Correction: v_off_ini = -75.812 uV, v_off_fin = -66.277 uV 12:48:07 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 22.578 uVs 12:48:07 Integrated Voltage Measurement... 12:48:16 Voltage Offset Correction: v_off_ini = -60.434 uV, v_off_fin = -65.504 uV 12:48:16 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -19.677 uVs 12:48:16 Integrated Voltage Measurement... 12:48:25 Voltage Offset Correction: v_off_ini = -74.539 uV, v_off_fin = -69.395 uV 12:48:25 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 37.409 uVs 12:48:25 Integrated Voltage Measurement... 12:48:34 Voltage Offset Correction: v_off_ini = -61.391 uV, v_off_fin = -63.243 uV 12:48:34 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -19.309 uVs 12:48:34 Integrated Voltage Measurement... 12:48:43 Voltage Offset Correction: v_off_ini = -73.729 uV, v_off_fin = -65.553 uV 12:48:43 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 29.202 uVs 12:48:48 I1X Average and RMS of Measurements VT = -27.787 +- 2.630 uVs I1X = 5.558 +- 0.526 uTm 12:48:50 Wire Position: X0 = 0.0000 mm, Y0 = 0.5000 mm 12:48:54 Integrated Voltage Measurement... 12:49:03 Voltage Offset Correction: v_off_ini = -72.829 uV, v_off_fin = -71.184 uV 12:49:03 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -133.857 uVs 12:49:03 Integrated Voltage Measurement... 12:49:12 Voltage Offset Correction: v_off_ini = -60.946 uV, v_off_fin = -65.456 uV 12:49:12 Integrated Voltage Measurement dXYus = -0.5001 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 166.713 uVs 12:49:12 Integrated Voltage Measurement... 12:49:21 Voltage Offset Correction: v_off_ini = -65.143 uV, v_off_fin = -89.999 uV 12:49:21 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -148.708 uVs 12:49:21 Integrated Voltage Measurement... 12:49:30 Voltage Offset Correction: v_off_ini = -71.645 uV, v_off_fin = -64.138 uV 12:49:30 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 165.049 uVs 12:49:30 Integrated Voltage Measurement... 12:49:39 Voltage Offset Correction: v_off_ini = -72.829 uV, v_off_fin = -88.816 uV 12:49:39 Integrated Voltage Measurement dXYus = 0.5001 mm, dXYds = 0.5001 mm dXY = 0.5001 mm VT = -162.432 uVs 12:49:39 Integrated Voltage Measurement... 12:49:48 Voltage Offset Correction: v_off_ini = -73.860 uV, v_off_fin = -65.866 uV 12:49:48 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 136.826 uVs 12:49:53 I1Y Average and RMS of Measurements VT = -152.264 +- 3.274 uVs I1Y = -30.451 +- 0.656 uTm 12:49:55 Wire Position: X0 = 0.0000 mm, Y0 = 0.5000 mm 12:49:58 Integrated Voltage Measurement... 12:50:07 Voltage Offset Correction: v_off_ini = -70.252 uV, v_off_fin = -61.475 uV 12:50:07 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -51.270 uVs 12:50:07 Integrated Voltage Measurement... 12:50:16 Voltage Offset Correction: v_off_ini = -75.247 uV, v_off_fin = -81.798 uV 12:50:16 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 22.599 uVs 12:50:16 Integrated Voltage Measurement... 12:50:25 Voltage Offset Correction: v_off_ini = -76.228 uV, v_off_fin = -53.307 uV 12:50:25 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -49.396 uVs 12:50:25 Integrated Voltage Measurement... 12:50:34 Voltage Offset Correction: v_off_ini = -90.307 uV, v_off_fin = -79.899 uV 12:50:34 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 4.759 uVs 12:50:34 Integrated Voltage Measurement... 12:50:43 Voltage Offset Correction: v_off_ini = -78.156 uV, v_off_fin = -66.300 uV 12:50:43 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -34.545 uVs 12:50:43 Integrated Voltage Measurement... 12:50:52 Voltage Offset Correction: v_off_ini = -68.126 uV, v_off_fin = -68.740 uV 12:50:52 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 27.152 uVs 12:50:54 I2X Average and RMS of Measurements VT = -31.620 +- 4.061 uVs I2X = 2.909 +- 0.374 uTm^2 12:50:57 Wire Position: X0 = 0.0000 mm, Y0 = 0.5000 mm 12:50:59 Integrated Voltage Measurement... 12:51:08 Voltage Offset Correction: v_off_ini = -67.820 uV, v_off_fin = -66.882 uV 12:51:08 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -191.402 uVs 12:51:08 Integrated Voltage Measurement... 12:51:17 Voltage Offset Correction: v_off_ini = -63.642 uV, v_off_fin = -71.139 uV 12:51:17 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 194.010 uVs 12:51:18 Integrated Voltage Measurement... 12:51:26 Voltage Offset Correction: v_off_ini = -78.168 uV, v_off_fin = -65.975 uV 12:51:27 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -176.589 uVs 12:51:27 Integrated Voltage Measurement... 12:51:36 Voltage Offset Correction: v_off_ini = -70.967 uV, v_off_fin = -71.206 uV 12:51:36 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 178.045 uVs 12:51:36 Integrated Voltage Measurement... 12:51:45 Voltage Offset Correction: v_off_ini = -59.755 uV, v_off_fin = -80.884 uV 12:51:45 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -161.088 uVs 12:51:45 Integrated Voltage Measurement... 12:51:54 Voltage Offset Correction: v_off_ini = -71.263 uV, v_off_fin = -74.885 uV 12:51:54 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 179.032 uVs 12:51:56 I2Y Average and RMS of Measurements VT = -180.028 +- 9.442 uVs I2Y = -16.563 +- 0.869 uTm^2 12:52:00 Wire Position: X0 = 0.0000 mm, Y0 = 1.0000 mm 12:52:04 Integrated Voltage Measurement... 12:52:13 Voltage Offset Correction: v_off_ini = -66.368 uV, v_off_fin = -70.541 uV 12:52:13 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -19.470 uVs 12:52:13 Integrated Voltage Measurement... 12:52:22 Voltage Offset Correction: v_off_ini = -78.153 uV, v_off_fin = -84.316 uV 12:52:22 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 27.799 uVs 12:52:22 Integrated Voltage Measurement... 12:52:31 Voltage Offset Correction: v_off_ini = -80.552 uV, v_off_fin = -65.732 uV 12:52:31 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -30.656 uVs 12:52:31 Integrated Voltage Measurement... 12:52:40 Voltage Offset Correction: v_off_ini = -67.270 uV, v_off_fin = -61.333 uV 12:52:40 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 30.817 uVs 12:52:40 Integrated Voltage Measurement... 12:52:49 Voltage Offset Correction: v_off_ini = -71.102 uV, v_off_fin = -69.462 uV 12:52:49 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -25.779 uVs 12:52:49 Integrated Voltage Measurement... 12:52:58 Voltage Offset Correction: v_off_ini = -71.924 uV, v_off_fin = -61.179 uV 12:52:58 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 32.622 uVs 12:53:03 I1X Average and RMS of Measurements VT = -27.857 +- 3.051 uVs I1X = 5.571 +- 0.610 uTm 12:53:05 Wire Position: X0 = 0.0000 mm, Y0 = 1.0000 mm 12:53:09 Integrated Voltage Measurement... 12:53:18 Voltage Offset Correction: v_off_ini = -65.245 uV, v_off_fin = -78.404 uV 12:53:18 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -166.934 uVs 12:53:18 Integrated Voltage Measurement... 12:53:27 Voltage Offset Correction: v_off_ini = -79.326 uV, v_off_fin = -64.186 uV 12:53:27 Integrated Voltage Measurement dXYus = -0.5001 mm, dXYds = -0.5001 mm dXY = -0.5001 mm VT = 166.770 uVs 12:53:27 Integrated Voltage Measurement... 12:53:36 Voltage Offset Correction: v_off_ini = -53.949 uV, v_off_fin = -77.606 uV 12:53:36 Integrated Voltage Measurement dXYus = 0.5001 mm, dXYds = 0.5001 mm dXY = 0.5001 mm VT = -168.430 uVs 12:53:36 Integrated Voltage Measurement... 12:53:45 Voltage Offset Correction: v_off_ini = -64.493 uV, v_off_fin = -58.225 uV 12:53:45 Integrated Voltage Measurement dXYus = -0.5001 mm, dXYds = -0.5001 mm dXY = -0.5001 mm VT = 166.795 uVs 12:53:45 Integrated Voltage Measurement... 12:53:54 Voltage Offset Correction: v_off_ini = -78.611 uV, v_off_fin = -83.238 uV 12:53:54 Integrated Voltage Measurement dXYus = 0.5001 mm, dXYds = 0.5002 mm dXY = 0.5001 mm VT = -175.262 uVs 12:53:54 Integrated Voltage Measurement... 12:54:03 Voltage Offset Correction: v_off_ini = -60.431 uV, v_off_fin = -51.567 uV 12:54:03 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 146.011 uVs 12:54:08 I1Y Average and RMS of Measurements VT = -165.034 +- 3.125 uVs I1Y = -33.001 +- 0.626 uTm 12:54:10 Wire Position: X0 = 0.0000 mm, Y0 = 1.0000 mm 12:54:13 Integrated Voltage Measurement... 12:54:22 Voltage Offset Correction: v_off_ini = -68.966 uV, v_off_fin = -66.267 uV 12:54:22 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -18.077 uVs 12:54:22 Integrated Voltage Measurement... 12:54:31 Voltage Offset Correction: v_off_ini = -55.774 uV, v_off_fin = -72.274 uV 12:54:31 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 28.394 uVs 12:54:31 Integrated Voltage Measurement... 12:54:40 Voltage Offset Correction: v_off_ini = -63.572 uV, v_off_fin = -66.472 uV 12:54:40 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -24.978 uVs 12:54:40 Integrated Voltage Measurement... 12:54:49 Voltage Offset Correction: v_off_ini = -80.062 uV, v_off_fin = -88.056 uV 12:54:49 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 37.176 uVs 12:54:49 Integrated Voltage Measurement... 12:54:58 Voltage Offset Correction: v_off_ini = -61.042 uV, v_off_fin = -59.590 uV 12:54:58 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -41.643 uVs 12:54:58 Integrated Voltage Measurement... 12:55:07 Voltage Offset Correction: v_off_ini = -66.040 uV, v_off_fin = -68.673 uV 12:55:07 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 43.991 uVs 12:55:09 I2X Average and RMS of Measurements VT = -32.377 +- 8.047 uVs I2X = 2.979 +- 0.740 uTm^2 12:55:12 Wire Position: X0 = 0.0000 mm, Y0 = 1.0000 mm 12:55:14 Integrated Voltage Measurement... 12:55:23 Voltage Offset Correction: v_off_ini = -85.501 uV, v_off_fin = -78.064 uV 12:55:23 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -155.442 uVs 12:55:24 Integrated Voltage Measurement... 12:55:32 Voltage Offset Correction: v_off_ini = -68.631 uV, v_off_fin = -62.124 uV 12:55:33 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 191.099 uVs 12:55:33 Integrated Voltage Measurement... 12:55:42 Voltage Offset Correction: v_off_ini = -60.769 uV, v_off_fin = -76.188 uV 12:55:42 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = -202.147 uVs 12:55:42 Integrated Voltage Measurement... 12:55:51 Voltage Offset Correction: v_off_ini = -73.146 uV, v_off_fin = -74.069 uV 12:55:51 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 168.725 uVs 12:55:51 Integrated Voltage Measurement... 12:56:00 Voltage Offset Correction: v_off_ini = -84.794 uV, v_off_fin = -87.011 uV 12:56:00 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -167.187 uVs 12:56:00 Integrated Voltage Measurement... 12:56:09 Voltage Offset Correction: v_off_ini = -75.328 uV, v_off_fin = -74.503 uV 12:56:09 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = 175.556 uVs 12:56:11 I2Y Average and RMS of Measurements VT = -176.693 +- 6.231 uVs I2Y = -16.256 +- 0.573 uTm^2 12:56:27 Wire Position: X0 = -2.0000 mm, Y0 = 0.0000 mm 12:56:31 Integrated Voltage Measurement... 12:56:40 Voltage Offset Correction: v_off_ini = -73.681 uV, v_off_fin = -61.813 uV 12:56:40 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -18.507 uVs 12:56:40 Integrated Voltage Measurement... 12:56:49 Voltage Offset Correction: v_off_ini = -69.815 uV, v_off_fin = -70.780 uV 12:56:49 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -2.045 uVs 12:56:49 Integrated Voltage Measurement... 12:56:58 Voltage Offset Correction: v_off_ini = -75.425 uV, v_off_fin = -74.330 uV 12:56:58 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -32.100 uVs 12:56:58 Integrated Voltage Measurement... 12:57:07 Voltage Offset Correction: v_off_ini = -71.600 uV, v_off_fin = -72.851 uV 12:57:07 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 4.995 uVs 12:57:07 Integrated Voltage Measurement... 12:57:16 Voltage Offset Correction: v_off_ini = -70.454 uV, v_off_fin = -68.847 uV 12:57:16 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -29.644 uVs 12:57:16 Integrated Voltage Measurement... 12:57:25 Voltage Offset Correction: v_off_ini = -72.629 uV, v_off_fin = -77.918 uV 12:57:25 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 3.689 uVs 12:57:30 I1X Average and RMS of Measurements VT = -14.482 +- 4.486 uVs I1X = 2.896 +- 0.897 uTm 12:57:32 Wire Position: X0 = -2.0000 mm, Y0 = 0.0000 mm 12:57:36 Integrated Voltage Measurement... 12:57:45 Voltage Offset Correction: v_off_ini = -83.880 uV, v_off_fin = -81.892 uV 12:57:45 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.4999 mm dXY = 0.4999 mm VT = -166.492 uVs 12:57:45 Integrated Voltage Measurement... 12:57:54 Voltage Offset Correction: v_off_ini = -73.741 uV, v_off_fin = -73.628 uV 12:57:54 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 136.052 uVs 12:57:54 Integrated Voltage Measurement... 12:58:03 Voltage Offset Correction: v_off_ini = -75.708 uV, v_off_fin = -86.576 uV 12:58:03 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -150.803 uVs 12:58:03 Integrated Voltage Measurement... 12:58:12 Voltage Offset Correction: v_off_ini = -68.823 uV, v_off_fin = -65.552 uV 12:58:12 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 152.940 uVs 12:58:12 Integrated Voltage Measurement... 12:58:21 Voltage Offset Correction: v_off_ini = -81.359 uV, v_off_fin = -83.650 uV 12:58:21 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -161.301 uVs 12:58:21 Integrated Voltage Measurement... 12:58:30 Voltage Offset Correction: v_off_ini = -86.003 uV, v_off_fin = -71.652 uV 12:58:30 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 159.972 uVs 12:58:35 I1Y Average and RMS of Measurements VT = -154.593 +- 4.280 uVs I1Y = -30.920 +- 0.855 uTm 12:58:40 Wire Position: X0 = -2.0000 mm, Y0 = 0.0000 mm 12:58:43 Integrated Voltage Measurement... 12:58:51 Voltage Offset Correction: v_off_ini = -51.372 uV, v_off_fin = -56.829 uV 12:58:52 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 4.671 uVs 12:58:52 Integrated Voltage Measurement... 12:59:01 Voltage Offset Correction: v_off_ini = -46.384 uV, v_off_fin = -71.992 uV 12:59:01 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 18.691 uVs 12:59:01 Integrated Voltage Measurement... 12:59:10 Voltage Offset Correction: v_off_ini = -69.981 uV, v_off_fin = -60.477 uV 12:59:10 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -7.447 uVs 12:59:10 Integrated Voltage Measurement... 12:59:19 Voltage Offset Correction: v_off_ini = -51.632 uV, v_off_fin = -75.450 uV 12:59:19 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 19.074 uVs 12:59:19 Integrated Voltage Measurement... 12:59:28 Voltage Offset Correction: v_off_ini = -64.052 uV, v_off_fin = -58.284 uV 12:59:28 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -15.486 uVs 12:59:28 Integrated Voltage Measurement... 12:59:37 Voltage Offset Correction: v_off_ini = -63.163 uV, v_off_fin = -63.600 uV 12:59:37 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 9.710 uVs 12:59:39 I2X Average and RMS of Measurements VT = -10.956 +- 2.803 uVs I2X = 1.008 +- 0.258 uTm^2 12:59:44 Wire Position: X0 = -2.0000 mm, Y0 = 0.0000 mm 12:59:47 Integrated Voltage Measurement... 12:59:56 Voltage Offset Correction: v_off_ini = -85.045 uV, v_off_fin = -62.405 uV 12:59:56 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -171.367 uVs 12:59:56 Integrated Voltage Measurement... 13:00:05 Voltage Offset Correction: v_off_ini = -51.414 uV, v_off_fin = -79.773 uV 13:00:05 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 188.801 uVs 13:00:05 Integrated Voltage Measurement... 13:00:14 Voltage Offset Correction: v_off_ini = -65.058 uV, v_off_fin = -68.021 uV 13:00:14 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -187.686 uVs 13:00:14 Integrated Voltage Measurement... 13:00:23 Voltage Offset Correction: v_off_ini = -79.143 uV, v_off_fin = -71.659 uV 13:00:23 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 174.411 uVs 13:00:23 Integrated Voltage Measurement... 13:00:32 Voltage Offset Correction: v_off_ini = -60.973 uV, v_off_fin = -73.983 uV 13:00:32 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -166.511 uVs 13:00:32 Integrated Voltage Measurement... 13:00:41 Voltage Offset Correction: v_off_ini = -52.064 uV, v_off_fin = -56.908 uV 13:00:41 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 158.654 uVs 13:00:43 I2Y Average and RMS of Measurements VT = -174.572 +- 8.487 uVs I2Y = -16.061 +- 0.781 uTm^2 13:00:48 Wire Position: X0 = -1.0000 mm, Y0 = 0.0000 mm 13:00:52 Integrated Voltage Measurement... 13:01:01 Voltage Offset Correction: v_off_ini = -62.053 uV, v_off_fin = -74.869 uV 13:01:01 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -21.192 uVs 13:01:01 Integrated Voltage Measurement... 13:01:10 Voltage Offset Correction: v_off_ini = -58.425 uV, v_off_fin = -74.048 uV 13:01:10 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 24.733 uVs 13:01:10 Integrated Voltage Measurement... 13:01:19 Voltage Offset Correction: v_off_ini = -66.080 uV, v_off_fin = -67.474 uV 13:01:19 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -21.817 uVs 13:01:19 Integrated Voltage Measurement... 13:01:28 Voltage Offset Correction: v_off_ini = -57.075 uV, v_off_fin = -62.158 uV 13:01:28 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 15.526 uVs 13:01:28 Integrated Voltage Measurement... 13:01:37 Voltage Offset Correction: v_off_ini = -68.019 uV, v_off_fin = -79.155 uV 13:01:37 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -13.616 uVs 13:01:37 Integrated Voltage Measurement... 13:01:46 Voltage Offset Correction: v_off_ini = -63.055 uV, v_off_fin = -62.242 uV 13:01:46 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 27.818 uVs 13:01:51 I1X Average and RMS of Measurements VT = -20.784 +- 1.753 uVs I1X = 4.157 +- 0.351 uTm 13:01:53 Wire Position: X0 = -1.0000 mm, Y0 = 0.0000 mm 13:01:57 Integrated Voltage Measurement... 13:02:06 Voltage Offset Correction: v_off_ini = -93.166 uV, v_off_fin = -87.776 uV 13:02:06 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -144.311 uVs 13:02:06 Integrated Voltage Measurement... 13:02:15 Voltage Offset Correction: v_off_ini = -71.825 uV, v_off_fin = -64.251 uV 13:02:15 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.4999 mm dXY = -0.4999 mm VT = 149.299 uVs 13:02:15 Integrated Voltage Measurement... 13:02:24 Voltage Offset Correction: v_off_ini = -55.332 uV, v_off_fin = -68.361 uV 13:02:24 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.4999 mm dXY = 0.4999 mm VT = -154.727 uVs 13:02:24 Integrated Voltage Measurement... 13:02:33 Voltage Offset Correction: v_off_ini = -67.567 uV, v_off_fin = -64.583 uV 13:02:33 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 151.832 uVs 13:02:33 Integrated Voltage Measurement... 13:02:42 Voltage Offset Correction: v_off_ini = -61.810 uV, v_off_fin = -66.769 uV 13:02:42 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -159.815 uVs 13:02:42 Integrated Voltage Measurement... 13:02:51 Voltage Offset Correction: v_off_ini = -63.473 uV, v_off_fin = -54.518 uV 13:02:51 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 145.258 uVs 13:02:56 I1Y Average and RMS of Measurements VT = -150.874 +- 2.893 uVs I1Y = -30.177 +- 0.578 uTm 13:03:01 Wire Position: X0 = -1.0000 mm, Y0 = 0.0000 mm 13:03:03 Integrated Voltage Measurement... 13:03:12 Voltage Offset Correction: v_off_ini = -43.897 uV, v_off_fin = -58.338 uV 13:03:12 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -27.934 uVs 13:03:12 Integrated Voltage Measurement... 13:03:21 Voltage Offset Correction: v_off_ini = -61.231 uV, v_off_fin = -60.036 uV 13:03:21 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 25.344 uVs 13:03:21 Integrated Voltage Measurement... 13:03:30 Voltage Offset Correction: v_off_ini = -54.465 uV, v_off_fin = -54.946 uV 13:03:30 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -34.379 uVs 13:03:30 Integrated Voltage Measurement... 13:03:39 Voltage Offset Correction: v_off_ini = -51.959 uV, v_off_fin = -59.297 uV 13:03:39 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 41.515 uVs 13:03:40 Integrated Voltage Measurement... 13:03:49 Voltage Offset Correction: v_off_ini = -71.728 uV, v_off_fin = -56.792 uV 13:03:49 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -40.290 uVs 13:03:49 Integrated Voltage Measurement... 13:03:58 Voltage Offset Correction: v_off_ini = -73.051 uV, v_off_fin = -75.072 uV 13:03:58 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 36.305 uVs 13:04:00 I2X Average and RMS of Measurements VT = -34.294 +- 5.415 uVs I2X = 3.155 +- 0.498 uTm^2 13:04:05 Wire Position: X0 = -1.0000 mm, Y0 = 0.0000 mm 13:04:08 Integrated Voltage Measurement... 13:04:17 Voltage Offset Correction: v_off_ini = -74.351 uV, v_off_fin = -70.295 uV 13:04:17 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -162.242 uVs 13:04:17 Integrated Voltage Measurement... 13:04:26 Voltage Offset Correction: v_off_ini = -63.929 uV, v_off_fin = -67.392 uV 13:04:26 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 157.007 uVs 13:04:26 Integrated Voltage Measurement... 13:04:35 Voltage Offset Correction: v_off_ini = -61.081 uV, v_off_fin = -72.259 uV 13:04:35 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -188.775 uVs 13:04:35 Integrated Voltage Measurement... 13:04:44 Voltage Offset Correction: v_off_ini = -77.248 uV, v_off_fin = -68.700 uV 13:04:44 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 161.115 uVs 13:04:44 Integrated Voltage Measurement... 13:04:53 Voltage Offset Correction: v_off_ini = -66.795 uV, v_off_fin = -83.601 uV 13:04:53 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = -155.891 uVs 13:04:53 Integrated Voltage Measurement... 13:05:02 Voltage Offset Correction: v_off_ini = -82.398 uV, v_off_fin = -77.662 uV 13:05:02 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 178.003 uVs 13:05:04 I2Y Average and RMS of Measurements VT = -167.172 +- 6.256 uVs I2Y = -15.380 +- 0.576 uTm^2 13:05:09 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 13:05:13 Integrated Voltage Measurement... 13:05:22 Voltage Offset Correction: v_off_ini = -82.324 uV, v_off_fin = -69.674 uV 13:05:22 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -54.056 uVs 13:05:22 Integrated Voltage Measurement... 13:05:31 Voltage Offset Correction: v_off_ini = -67.509 uV, v_off_fin = -72.410 uV 13:05:31 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 41.781 uVs 13:05:31 Integrated Voltage Measurement... 13:05:40 Voltage Offset Correction: v_off_ini = -78.771 uV, v_off_fin = -80.676 uV 13:05:40 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -38.528 uVs 13:05:40 Integrated Voltage Measurement... 13:05:49 Voltage Offset Correction: v_off_ini = -77.276 uV, v_off_fin = -80.619 uV 13:05:49 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 4.085 uVs 13:05:49 Integrated Voltage Measurement... 13:05:58 Voltage Offset Correction: v_off_ini = -71.112 uV, v_off_fin = -83.886 uV 13:05:58 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -54.802 uVs 13:05:58 Integrated Voltage Measurement... 13:06:07 Voltage Offset Correction: v_off_ini = -68.068 uV, v_off_fin = -92.387 uV 13:06:07 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 24.011 uVs 13:06:12 I1X Average and RMS of Measurements VT = -36.211 +- 11.097 uVs I1X = 7.242 +- 2.220 uTm 13:06:14 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 13:06:18 Integrated Voltage Measurement... 13:06:27 Voltage Offset Correction: v_off_ini = -87.997 uV, v_off_fin = -88.674 uV 13:06:27 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -152.464 uVs 13:06:27 Integrated Voltage Measurement... 13:06:36 Voltage Offset Correction: v_off_ini = -82.590 uV, v_off_fin = -66.323 uV 13:06:36 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 156.978 uVs 13:06:36 Integrated Voltage Measurement... 13:06:45 Voltage Offset Correction: v_off_ini = -74.723 uV, v_off_fin = -71.409 uV 13:06:45 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -154.998 uVs 13:06:45 Integrated Voltage Measurement... 13:06:54 Voltage Offset Correction: v_off_ini = -84.092 uV, v_off_fin = -67.692 uV 13:06:54 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 155.299 uVs 13:06:54 Integrated Voltage Measurement... 13:07:03 Voltage Offset Correction: v_off_ini = -67.260 uV, v_off_fin = -85.675 uV 13:07:03 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -153.424 uVs 13:07:03 Integrated Voltage Measurement... 13:07:12 Voltage Offset Correction: v_off_ini = -76.412 uV, v_off_fin = -79.994 uV 13:07:12 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 148.848 uVs 13:07:17 I1Y Average and RMS of Measurements VT = -153.668 +- 1.799 uVs I1Y = -30.734 +- 0.360 uTm 13:07:22 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 13:07:24 Integrated Voltage Measurement... 13:07:33 Voltage Offset Correction: v_off_ini = -82.344 uV, v_off_fin = -88.891 uV 13:07:33 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -42.522 uVs 13:07:33 Integrated Voltage Measurement... 13:07:42 Voltage Offset Correction: v_off_ini = -80.308 uV, v_off_fin = -91.926 uV 13:07:42 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 61.139 uVs 13:07:42 Integrated Voltage Measurement... 13:07:51 Voltage Offset Correction: v_off_ini = -80.517 uV, v_off_fin = -77.059 uV 13:07:51 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -31.372 uVs 13:07:51 Integrated Voltage Measurement... 13:08:00 Voltage Offset Correction: v_off_ini = -90.493 uV, v_off_fin = -84.369 uV 13:08:00 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 52.269 uVs 13:08:00 Integrated Voltage Measurement... 13:08:09 Voltage Offset Correction: v_off_ini = -87.008 uV, v_off_fin = -66.220 uV 13:08:09 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -39.038 uVs 13:08:09 Integrated Voltage Measurement... 13:08:18 Voltage Offset Correction: v_off_ini = -74.864 uV, v_off_fin = -81.102 uV 13:08:18 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 37.472 uVs 13:08:21 I2X Average and RMS of Measurements VT = -43.969 +- 5.747 uVs I2X = 4.045 +- 0.529 uTm^2 13:08:26 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 13:08:29 Integrated Voltage Measurement... 13:08:38 Voltage Offset Correction: v_off_ini = -75.173 uV, v_off_fin = -85.759 uV 13:08:38 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -176.325 uVs 13:08:38 Integrated Voltage Measurement... 13:08:47 Voltage Offset Correction: v_off_ini = -93.100 uV, v_off_fin = -90.517 uV 13:08:47 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 181.312 uVs 13:08:47 Integrated Voltage Measurement... 13:08:56 Voltage Offset Correction: v_off_ini = -90.014 uV, v_off_fin = -86.741 uV 13:08:56 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -180.166 uVs 13:08:56 Integrated Voltage Measurement... 13:09:05 Voltage Offset Correction: v_off_ini = -83.868 uV, v_off_fin = -83.285 uV 13:09:05 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 163.449 uVs 13:09:05 Integrated Voltage Measurement... 13:09:14 Voltage Offset Correction: v_off_ini = -74.964 uV, v_off_fin = -76.861 uV 13:09:14 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = -165.982 uVs 13:09:14 Integrated Voltage Measurement... 13:09:23 Voltage Offset Correction: v_off_ini = -73.343 uV, v_off_fin = -74.255 uV 13:09:23 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 171.693 uVs 13:09:25 I2Y Average and RMS of Measurements VT = -173.154 +- 4.184 uVs I2Y = -15.930 +- 0.385 uTm^2 13:09:30 Wire Position: X0 = 1.0000 mm, Y0 = 0.0000 mm 13:09:34 Integrated Voltage Measurement... 13:09:43 Voltage Offset Correction: v_off_ini = -71.065 uV, v_off_fin = -66.187 uV 13:09:43 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -37.645 uVs 13:09:43 Integrated Voltage Measurement... 13:09:52 Voltage Offset Correction: v_off_ini = -70.372 uV, v_off_fin = -66.976 uV 13:09:52 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 62.655 uVs 13:09:52 Integrated Voltage Measurement... 13:10:01 Voltage Offset Correction: v_off_ini = -69.743 uV, v_off_fin = -85.891 uV 13:10:01 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -12.279 uVs 13:10:01 Integrated Voltage Measurement... 13:10:10 Voltage Offset Correction: v_off_ini = -84.178 uV, v_off_fin = -71.250 uV 13:10:10 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 38.884 uVs 13:10:10 Integrated Voltage Measurement... 13:10:19 Voltage Offset Correction: v_off_ini = -70.242 uV, v_off_fin = -68.511 uV 13:10:19 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -42.220 uVs 13:10:19 Integrated Voltage Measurement... 13:10:28 Voltage Offset Correction: v_off_ini = -60.456 uV, v_off_fin = -57.775 uV 13:10:28 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 38.307 uVs 13:10:32 I1X Average and RMS of Measurements VT = -38.665 +- 10.094 uVs I1X = 7.733 +- 2.019 uTm 13:10:35 Wire Position: X0 = 1.0000 mm, Y0 = 0.0000 mm 13:10:39 Integrated Voltage Measurement... 13:10:48 Voltage Offset Correction: v_off_ini = -63.106 uV, v_off_fin = -76.004 uV 13:10:48 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -153.273 uVs 13:10:48 Integrated Voltage Measurement... 13:10:57 Voltage Offset Correction: v_off_ini = -58.990 uV, v_off_fin = -52.517 uV 13:10:57 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 139.426 uVs 13:10:57 Integrated Voltage Measurement... 13:11:06 Voltage Offset Correction: v_off_ini = -62.427 uV, v_off_fin = -76.682 uV 13:11:06 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -153.298 uVs 13:11:06 Integrated Voltage Measurement... 13:11:15 Voltage Offset Correction: v_off_ini = -63.616 uV, v_off_fin = -56.342 uV 13:11:15 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 159.731 uVs 13:11:15 Integrated Voltage Measurement... 13:11:24 Voltage Offset Correction: v_off_ini = -76.048 uV, v_off_fin = -70.282 uV 13:11:24 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -138.611 uVs 13:11:24 Integrated Voltage Measurement... 13:11:33 Voltage Offset Correction: v_off_ini = -60.267 uV, v_off_fin = -62.781 uV 13:11:33 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 138.650 uVs 13:11:37 I1Y Average and RMS of Measurements VT = -147.165 +- 7.324 uVs I1Y = -29.433 +- 1.465 uTm 13:11:43 Wire Position: X0 = 1.0000 mm, Y0 = 0.0000 mm 13:11:45 Integrated Voltage Measurement... 13:11:54 Voltage Offset Correction: v_off_ini = -47.335 uV, v_off_fin = -57.919 uV 13:11:54 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -44.279 uVs 13:11:54 Integrated Voltage Measurement... 13:12:03 Voltage Offset Correction: v_off_ini = -76.459 uV, v_off_fin = -68.679 uV 13:12:03 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 42.191 uVs 13:12:03 Integrated Voltage Measurement... 13:12:12 Voltage Offset Correction: v_off_ini = -62.378 uV, v_off_fin = -57.205 uV 13:12:12 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -47.247 uVs 13:12:12 Integrated Voltage Measurement... 13:12:21 Voltage Offset Correction: v_off_ini = -72.646 uV, v_off_fin = -76.558 uV 13:12:21 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 51.794 uVs 13:12:21 Integrated Voltage Measurement... 13:12:30 Voltage Offset Correction: v_off_ini = -72.478 uV, v_off_fin = -63.492 uV 13:12:30 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -37.548 uVs 13:12:30 Integrated Voltage Measurement... 13:12:39 Voltage Offset Correction: v_off_ini = -68.258 uV, v_off_fin = -81.038 uV 13:12:39 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 31.717 uVs 13:12:42 I2X Average and RMS of Measurements VT = -42.463 +- 6.103 uVs I2X = 3.907 +- 0.561 uTm^2 13:12:47 Wire Position: X0 = 1.0000 mm, Y0 = 0.0000 mm 13:12:50 Integrated Voltage Measurement... 13:12:58 Voltage Offset Correction: v_off_ini = -84.405 uV, v_off_fin = -74.497 uV 13:12:59 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -150.012 uVs 13:12:59 Integrated Voltage Measurement... 13:13:08 Voltage Offset Correction: v_off_ini = -70.809 uV, v_off_fin = -76.236 uV 13:13:08 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 172.326 uVs 13:13:08 Integrated Voltage Measurement... 13:13:17 Voltage Offset Correction: v_off_ini = -76.952 uV, v_off_fin = -82.131 uV 13:13:17 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -179.693 uVs 13:13:17 Integrated Voltage Measurement... 13:13:26 Voltage Offset Correction: v_off_ini = -95.349 uV, v_off_fin = -82.799 uV 13:13:26 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 172.555 uVs 13:13:26 Integrated Voltage Measurement... 13:13:35 Voltage Offset Correction: v_off_ini = -86.451 uV, v_off_fin = -84.577 uV 13:13:35 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -146.793 uVs 13:13:35 Integrated Voltage Measurement... 13:13:44 Voltage Offset Correction: v_off_ini = -91.230 uV, v_off_fin = -100.641 uV 13:13:44 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 142.476 uVs 13:13:46 I2Y Average and RMS of Measurements VT = -160.642 +- 12.861 uVs I2Y = -14.779 +- 1.183 uTm^2 13:13:51 Wire Position: X0 = 2.0000 mm, Y0 = 0.0000 mm 13:13:55 Integrated Voltage Measurement... 13:14:04 Voltage Offset Correction: v_off_ini = -107.369 uV, v_off_fin = -98.846 uV 13:14:04 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -56.072 uVs 13:14:04 Integrated Voltage Measurement... 13:14:13 Voltage Offset Correction: v_off_ini = -95.136 uV, v_off_fin = -117.359 uV 13:14:13 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 22.948 uVs 13:14:13 Integrated Voltage Measurement... 13:14:22 Voltage Offset Correction: v_off_ini = -93.126 uV, v_off_fin = -91.110 uV 13:14:22 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -57.102 uVs 13:14:22 Integrated Voltage Measurement... 13:14:31 Voltage Offset Correction: v_off_ini = -108.397 uV, v_off_fin = -91.357 uV 13:14:31 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 76.539 uVs 13:14:31 Integrated Voltage Measurement... 13:14:40 Voltage Offset Correction: v_off_ini = -101.280 uV, v_off_fin = -96.980 uV 13:14:40 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -39.597 uVs 13:14:40 Integrated Voltage Measurement... 13:14:49 Voltage Offset Correction: v_off_ini = -93.085 uV, v_off_fin = -95.343 uV 13:14:49 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 26.859 uVs 13:14:53 I1X Average and RMS of Measurements VT = -46.520 +- 14.582 uVs I1X = 9.304 +- 2.916 uTm 13:14:56 Wire Position: X0 = 2.0000 mm, Y0 = 0.0000 mm 13:15:00 Integrated Voltage Measurement... 13:15:09 Voltage Offset Correction: v_off_ini = -102.689 uV, v_off_fin = -106.497 uV 13:15:09 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -154.859 uVs 13:15:09 Integrated Voltage Measurement... 13:15:18 Voltage Offset Correction: v_off_ini = -93.956 uV, v_off_fin = -83.369 uV 13:15:18 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 155.964 uVs 13:15:18 Integrated Voltage Measurement... 13:15:27 Voltage Offset Correction: v_off_ini = -75.393 uV, v_off_fin = -82.737 uV 13:15:27 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -146.102 uVs 13:15:27 Integrated Voltage Measurement... 13:15:36 Voltage Offset Correction: v_off_ini = -86.171 uV, v_off_fin = -84.633 uV 13:15:36 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 160.763 uVs 13:15:36 Integrated Voltage Measurement... 13:15:45 Voltage Offset Correction: v_off_ini = -92.481 uV, v_off_fin = -87.220 uV 13:15:45 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -148.433 uVs 13:15:45 Integrated Voltage Measurement... 13:15:54 Voltage Offset Correction: v_off_ini = -71.900 uV, v_off_fin = -63.162 uV 13:15:54 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 170.050 uVs 13:15:58 I1Y Average and RMS of Measurements VT = -156.029 +- 2.411 uVs I1Y = -31.206 +- 0.482 uTm 13:16:03 Wire Position: X0 = 2.0000 mm, Y0 = 0.0000 mm 13:16:06 Integrated Voltage Measurement... 13:16:15 Voltage Offset Correction: v_off_ini = -89.939 uV, v_off_fin = -72.120 uV 13:16:15 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -63.075 uVs 13:16:15 Integrated Voltage Measurement... 13:16:24 Voltage Offset Correction: v_off_ini = -84.985 uV, v_off_fin = -89.203 uV 13:16:24 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 37.594 uVs 13:16:24 Integrated Voltage Measurement... 13:16:33 Voltage Offset Correction: v_off_ini = -80.700 uV, v_off_fin = -74.293 uV 13:16:33 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -49.624 uVs 13:16:33 Integrated Voltage Measurement... 13:16:42 Voltage Offset Correction: v_off_ini = -80.887 uV, v_off_fin = -92.673 uV 13:16:42 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 45.327 uVs 13:16:42 Integrated Voltage Measurement... 13:16:51 Voltage Offset Correction: v_off_ini = -82.499 uV, v_off_fin = -74.269 uV 13:16:51 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -33.268 uVs 13:16:51 Integrated Voltage Measurement... 13:17:00 Voltage Offset Correction: v_off_ini = -74.793 uV, v_off_fin = -87.101 uV 13:17:00 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 40.004 uVs 13:17:03 I2X Average and RMS of Measurements VT = -44.815 +- 5.900 uVs I2X = 4.123 +- 0.543 uTm^2 13:17:08 Wire Position: X0 = 2.0000 mm, Y0 = 0.0000 mm 13:17:10 Integrated Voltage Measurement... 13:17:19 Voltage Offset Correction: v_off_ini = -73.977 uV, v_off_fin = -72.226 uV 13:17:19 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -170.726 uVs 13:17:19 Integrated Voltage Measurement... 13:17:28 Voltage Offset Correction: v_off_ini = -71.528 uV, v_off_fin = -74.588 uV 13:17:28 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 156.237 uVs 13:17:28 Integrated Voltage Measurement... 13:17:37 Voltage Offset Correction: v_off_ini = -70.150 uV, v_off_fin = -62.172 uV 13:17:37 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -179.263 uVs 13:17:37 Integrated Voltage Measurement... 13:17:46 Voltage Offset Correction: v_off_ini = -72.619 uV, v_off_fin = -65.458 uV 13:17:46 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 178.932 uVs 13:17:47 Integrated Voltage Measurement... 13:17:55 Voltage Offset Correction: v_off_ini = -74.456 uV, v_off_fin = -77.253 uV 13:17:56 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -200.199 uVs 13:17:56 Integrated Voltage Measurement... 13:18:05 Voltage Offset Correction: v_off_ini = -82.670 uV, v_off_fin = -77.313 uV 13:18:05 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 170.665 uVs 13:18:07 I2Y Average and RMS of Measurements VT = -176.004 +- 9.224 uVs I2Y = -16.192 +- 0.849 uTm^2 13:18:15 Wire Position: X0 = 0.0000 mm, Y0 = -1.0000 mm 13:18:19 Integrated Voltage Measurement... 13:18:28 Voltage Offset Correction: v_off_ini = -85.496 uV, v_off_fin = -83.398 uV 13:18:28 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -26.809 uVs 13:18:28 Integrated Voltage Measurement... 13:18:37 Voltage Offset Correction: v_off_ini = -74.092 uV, v_off_fin = -66.548 uV 13:18:37 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 25.482 uVs 13:18:37 Integrated Voltage Measurement... 13:18:46 Voltage Offset Correction: v_off_ini = -66.081 uV, v_off_fin = -76.143 uV 13:18:46 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -45.814 uVs 13:18:46 Integrated Voltage Measurement... 13:18:55 Voltage Offset Correction: v_off_ini = -81.831 uV, v_off_fin = -79.519 uV 13:18:55 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 19.112 uVs 13:18:55 Integrated Voltage Measurement... 13:19:04 Voltage Offset Correction: v_off_ini = -85.123 uV, v_off_fin = -67.557 uV 13:19:04 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -25.655 uVs 13:19:04 Integrated Voltage Measurement... 13:19:13 Voltage Offset Correction: v_off_ini = -76.742 uV, v_off_fin = -80.451 uV 13:19:13 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 42.347 uVs 13:19:17 I1X Average and RMS of Measurements VT = -30.870 +- 3.399 uVs I1X = 6.174 +- 0.680 uTm 13:19:20 Wire Position: X0 = 0.0000 mm, Y0 = -1.0000 mm 13:19:24 Integrated Voltage Measurement... 13:19:33 Voltage Offset Correction: v_off_ini = -103.058 uV, v_off_fin = -91.445 uV 13:19:33 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -155.652 uVs 13:19:33 Integrated Voltage Measurement... 13:19:42 Voltage Offset Correction: v_off_ini = -85.014 uV, v_off_fin = -86.642 uV 13:19:42 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 149.448 uVs 13:19:42 Integrated Voltage Measurement... 13:19:51 Voltage Offset Correction: v_off_ini = -81.462 uV, v_off_fin = -84.331 uV 13:19:51 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -163.596 uVs 13:19:51 Integrated Voltage Measurement... 13:20:00 Voltage Offset Correction: v_off_ini = -86.881 uV, v_off_fin = -74.788 uV 13:20:00 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 146.933 uVs 13:20:00 Integrated Voltage Measurement... 13:20:09 Voltage Offset Correction: v_off_ini = -86.323 uV, v_off_fin = -86.868 uV 13:20:09 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -154.751 uVs 13:20:09 Integrated Voltage Measurement... 13:20:18 Voltage Offset Correction: v_off_ini = -81.002 uV, v_off_fin = -68.034 uV 13:20:18 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 145.839 uVs 13:20:22 I1Y Average and RMS of Measurements VT = -152.703 +- 2.032 uVs I1Y = -30.541 +- 0.406 uTm 13:20:25 Wire Position: X0 = 0.0000 mm, Y0 = -1.0000 mm 13:20:27 Integrated Voltage Measurement... 13:20:36 Voltage Offset Correction: v_off_ini = -89.020 uV, v_off_fin = -81.855 uV 13:20:36 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -35.704 uVs 13:20:36 Integrated Voltage Measurement... 13:20:45 Voltage Offset Correction: v_off_ini = -106.724 uV, v_off_fin = -92.957 uV 13:20:45 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 23.279 uVs 13:20:45 Integrated Voltage Measurement... 13:20:54 Voltage Offset Correction: v_off_ini = -98.301 uV, v_off_fin = -80.096 uV 13:20:54 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -26.970 uVs 13:20:54 Integrated Voltage Measurement... 13:21:03 Voltage Offset Correction: v_off_ini = -84.120 uV, v_off_fin = -95.964 uV 13:21:04 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 24.374 uVs 13:21:04 Integrated Voltage Measurement... 13:21:13 Voltage Offset Correction: v_off_ini = -83.626 uV, v_off_fin = -85.468 uV 13:21:13 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -40.577 uVs 13:21:13 Integrated Voltage Measurement... 13:21:22 Voltage Offset Correction: v_off_ini = -75.104 uV, v_off_fin = -86.998 uV 13:21:22 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 34.366 uVs 13:21:24 I2X Average and RMS of Measurements VT = -30.878 +- 4.916 uVs I2X = 2.841 +- 0.452 uTm^2 13:21:26 Wire Position: X0 = 0.0000 mm, Y0 = -1.0000 mm 13:21:29 Integrated Voltage Measurement... 13:21:38 Voltage Offset Correction: v_off_ini = -101.102 uV, v_off_fin = -85.654 uV 13:21:38 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -172.001 uVs 13:21:38 Integrated Voltage Measurement... 13:21:47 Voltage Offset Correction: v_off_ini = -75.216 uV, v_off_fin = -75.767 uV 13:21:47 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 158.762 uVs 13:21:47 Integrated Voltage Measurement... 13:21:56 Voltage Offset Correction: v_off_ini = -81.304 uV, v_off_fin = -88.053 uV 13:21:56 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -176.952 uVs 13:21:56 Integrated Voltage Measurement... 13:22:05 Voltage Offset Correction: v_off_ini = -79.890 uV, v_off_fin = -78.308 uV 13:22:05 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 149.852 uVs 13:22:05 Integrated Voltage Measurement... 13:22:14 Voltage Offset Correction: v_off_ini = -85.584 uV, v_off_fin = -87.917 uV 13:22:14 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -154.166 uVs 13:22:14 Integrated Voltage Measurement... 13:22:23 Voltage Offset Correction: v_off_ini = -70.628 uV, v_off_fin = -66.249 uV 13:22:23 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 147.426 uVs 13:22:26 I2Y Average and RMS of Measurements VT = -159.860 +- 6.460 uVs I2Y = -14.707 +- 0.594 uTm^2 13:22:30 Wire Position: X0 = 0.0000 mm, Y0 = -0.5000 mm 13:22:34 Integrated Voltage Measurement... 13:22:43 Voltage Offset Correction: v_off_ini = -87.719 uV, v_off_fin = -77.988 uV 13:22:43 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -26.225 uVs 13:22:43 Integrated Voltage Measurement... 13:22:52 Voltage Offset Correction: v_off_ini = -75.857 uV, v_off_fin = -73.719 uV 13:22:52 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 28.617 uVs 13:22:52 Integrated Voltage Measurement... 13:23:01 Voltage Offset Correction: v_off_ini = -87.322 uV, v_off_fin = -77.474 uV 13:23:01 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -18.999 uVs 13:23:01 Integrated Voltage Measurement... 13:23:10 Voltage Offset Correction: v_off_ini = -72.368 uV, v_off_fin = -94.028 uV 13:23:10 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 11.967 uVs 13:23:10 Integrated Voltage Measurement... 13:23:19 Voltage Offset Correction: v_off_ini = -79.168 uV, v_off_fin = -90.014 uV 13:23:19 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -19.593 uVs 13:23:19 Integrated Voltage Measurement... 13:23:28 Voltage Offset Correction: v_off_ini = -98.913 uV, v_off_fin = -93.951 uV 13:23:28 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 32.940 uVs 13:23:32 I1X Average and RMS of Measurements VT = -23.057 +- 5.376 uVs I1X = 4.611 +- 1.075 uTm 13:23:35 Wire Position: X0 = 0.0000 mm, Y0 = -0.5000 mm 13:23:39 Integrated Voltage Measurement... 13:23:48 Voltage Offset Correction: v_off_ini = -92.020 uV, v_off_fin = -97.795 uV 13:23:48 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -144.938 uVs 13:23:48 Integrated Voltage Measurement... 13:23:57 Voltage Offset Correction: v_off_ini = -84.032 uV, v_off_fin = -57.173 uV 13:23:57 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 163.776 uVs 13:23:57 Integrated Voltage Measurement... 13:24:06 Voltage Offset Correction: v_off_ini = -67.476 uV, v_off_fin = -83.858 uV 13:24:06 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -139.852 uVs 13:24:06 Integrated Voltage Measurement... 13:24:15 Voltage Offset Correction: v_off_ini = -92.605 uV, v_off_fin = -82.107 uV 13:24:15 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 166.217 uVs 13:24:15 Integrated Voltage Measurement... 13:24:24 Voltage Offset Correction: v_off_ini = -87.430 uV, v_off_fin = -106.527 uV 13:24:24 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -133.007 uVs 13:24:24 Integrated Voltage Measurement... 13:24:33 Voltage Offset Correction: v_off_ini = -109.658 uV, v_off_fin = -80.650 uV 13:24:33 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 150.719 uVs 13:24:37 I1Y Average and RMS of Measurements VT = -149.752 +- 5.604 uVs I1Y = -29.950 +- 1.121 uTm 13:24:40 Wire Position: X0 = 0.0000 mm, Y0 = -0.5000 mm 13:24:42 Integrated Voltage Measurement... 13:24:51 Voltage Offset Correction: v_off_ini = -91.816 uV, v_off_fin = -79.931 uV 13:24:51 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -19.332 uVs 13:24:51 Integrated Voltage Measurement... 13:25:00 Voltage Offset Correction: v_off_ini = -87.593 uV, v_off_fin = -97.689 uV 13:25:00 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 41.139 uVs 13:25:00 Integrated Voltage Measurement... 13:25:09 Voltage Offset Correction: v_off_ini = -89.084 uV, v_off_fin = -84.674 uV 13:25:10 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -25.128 uVs 13:25:10 Integrated Voltage Measurement... 13:25:19 Voltage Offset Correction: v_off_ini = -95.434 uV, v_off_fin = -97.789 uV 13:25:19 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 17.348 uVs 13:25:19 Integrated Voltage Measurement... 13:25:28 Voltage Offset Correction: v_off_ini = -90.480 uV, v_off_fin = -89.117 uV 13:25:28 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -12.556 uVs 13:25:28 Integrated Voltage Measurement... 13:25:37 Voltage Offset Correction: v_off_ini = -94.006 uV, v_off_fin = -89.058 uV 13:25:37 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 23.886 uVs 13:25:39 I2X Average and RMS of Measurements VT = -23.231 +- 5.103 uVs I2X = 2.137 +- 0.470 uTm^2 13:25:42 Wire Position: X0 = 0.0000 mm, Y0 = -0.5000 mm 13:25:44 Integrated Voltage Measurement... 13:25:53 Voltage Offset Correction: v_off_ini = -82.476 uV, v_off_fin = -88.540 uV 13:25:53 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -169.820 uVs 13:25:53 Integrated Voltage Measurement... 13:26:02 Voltage Offset Correction: v_off_ini = -90.889 uV, v_off_fin = -97.031 uV 13:26:02 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 184.938 uVs 13:26:02 Integrated Voltage Measurement... 13:26:11 Voltage Offset Correction: v_off_ini = -97.034 uV, v_off_fin = -110.489 uV 13:26:11 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -151.643 uVs 13:26:11 Integrated Voltage Measurement... 13:26:20 Voltage Offset Correction: v_off_ini = -86.829 uV, v_off_fin = -81.948 uV 13:26:20 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 177.163 uVs 13:26:20 Integrated Voltage Measurement... 13:26:29 Voltage Offset Correction: v_off_ini = -86.110 uV, v_off_fin = -77.909 uV 13:26:29 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -163.528 uVs 13:26:29 Integrated Voltage Measurement... 13:26:38 Voltage Offset Correction: v_off_ini = -92.952 uV, v_off_fin = -98.378 uV 13:26:38 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 154.552 uVs 13:26:41 I2Y Average and RMS of Measurements VT = -166.941 +- 7.699 uVs I2Y = -15.359 +- 0.708 uTm^2 13:26:45 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 13:26:49 Integrated Voltage Measurement... 13:26:58 Voltage Offset Correction: v_off_ini = -104.793 uV, v_off_fin = -96.937 uV 13:26:58 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -28.077 uVs 13:26:58 Integrated Voltage Measurement... 13:27:07 Voltage Offset Correction: v_off_ini = -92.334 uV, v_off_fin = -88.698 uV 13:27:07 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 11.354 uVs 13:27:07 Integrated Voltage Measurement... 13:27:16 Voltage Offset Correction: v_off_ini = -99.145 uV, v_off_fin = -94.124 uV 13:27:16 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -35.951 uVs 13:27:16 Integrated Voltage Measurement... 13:27:25 Voltage Offset Correction: v_off_ini = -97.316 uV, v_off_fin = -98.588 uV 13:27:25 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 17.369 uVs 13:27:25 Integrated Voltage Measurement... 13:27:34 Voltage Offset Correction: v_off_ini = -104.239 uV, v_off_fin = -108.615 uV 13:27:34 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -26.671 uVs 13:27:34 Integrated Voltage Measurement... 13:27:43 Voltage Offset Correction: v_off_ini = -108.759 uV, v_off_fin = -115.427 uV 13:27:43 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 46.365 uVs 13:27:47 I1X Average and RMS of Measurements VT = -27.631 +- 6.894 uVs I1X = 5.526 +- 1.379 uTm 13:27:50 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 13:27:54 Integrated Voltage Measurement... 13:28:03 Voltage Offset Correction: v_off_ini = -111.515 uV, v_off_fin = -132.971 uV 13:28:03 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -136.233 uVs 13:28:03 Integrated Voltage Measurement... 13:28:12 Voltage Offset Correction: v_off_ini = -122.830 uV, v_off_fin = -106.526 uV 13:28:12 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.4999 mm dXY = -0.4999 mm VT = 168.464 uVs 13:28:12 Integrated Voltage Measurement... 13:28:21 Voltage Offset Correction: v_off_ini = -99.681 uV, v_off_fin = -117.590 uV 13:28:21 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.4999 mm dXY = 0.4999 mm VT = -164.340 uVs 13:28:21 Integrated Voltage Measurement... 13:28:30 Voltage Offset Correction: v_off_ini = -112.242 uV, v_off_fin = -92.692 uV 13:28:30 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 158.076 uVs 13:28:30 Integrated Voltage Measurement... 13:28:39 Voltage Offset Correction: v_off_ini = -114.469 uV, v_off_fin = -123.909 uV 13:28:39 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.4999 mm dXY = 0.4999 mm VT = -150.993 uVs 13:28:39 Integrated Voltage Measurement... 13:28:48 Voltage Offset Correction: v_off_ini = -99.553 uV, v_off_fin = -79.811 uV 13:28:48 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 154.122 uVs 13:28:52 I1Y Average and RMS of Measurements VT = -155.371 +- 4.128 uVs I1Y = -31.077 +- 0.826 uTm 13:28:55 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 13:28:57 Integrated Voltage Measurement... 13:29:06 Voltage Offset Correction: v_off_ini = -100.699 uV, v_off_fin = -90.267 uV 13:29:06 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -42.668 uVs 13:29:06 Integrated Voltage Measurement... 13:29:15 Voltage Offset Correction: v_off_ini = -86.657 uV, v_off_fin = -93.305 uV 13:29:15 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 35.421 uVs 13:29:16 Integrated Voltage Measurement... 13:29:25 Voltage Offset Correction: v_off_ini = -87.935 uV, v_off_fin = -88.157 uV 13:29:25 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -33.256 uVs 13:29:25 Integrated Voltage Measurement... 13:29:34 Voltage Offset Correction: v_off_ini = -94.690 uV, v_off_fin = -100.905 uV 13:29:34 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 40.524 uVs 13:29:34 Integrated Voltage Measurement... 13:29:43 Voltage Offset Correction: v_off_ini = -91.475 uV, v_off_fin = -94.333 uV 13:29:43 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -47.580 uVs 13:29:43 Integrated Voltage Measurement... 13:29:52 Voltage Offset Correction: v_off_ini = -98.714 uV, v_off_fin = -99.936 uV 13:29:52 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 38.830 uVs 13:29:54 I2X Average and RMS of Measurements VT = -39.713 +- 2.621 uVs I2X = 3.654 +- 0.241 uTm^2 13:29:57 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 13:29:59 Integrated Voltage Measurement... 13:30:08 Voltage Offset Correction: v_off_ini = -90.299 uV, v_off_fin = -121.759 uV 13:30:08 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -168.249 uVs 13:30:08 Integrated Voltage Measurement... 13:30:17 Voltage Offset Correction: v_off_ini = -104.147 uV, v_off_fin = -97.523 uV 13:30:17 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 165.618 uVs 13:30:17 Integrated Voltage Measurement... 13:30:26 Voltage Offset Correction: v_off_ini = -88.272 uV, v_off_fin = -88.754 uV 13:30:26 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = -178.383 uVs 13:30:26 Integrated Voltage Measurement... 13:30:35 Voltage Offset Correction: v_off_ini = -92.803 uV, v_off_fin = -81.330 uV 13:30:35 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = 166.792 uVs 13:30:35 Integrated Voltage Measurement... 13:30:44 Voltage Offset Correction: v_off_ini = -100.134 uV, v_off_fin = -93.544 uV 13:30:44 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -161.841 uVs 13:30:44 Integrated Voltage Measurement... 13:30:53 Voltage Offset Correction: v_off_ini = -98.743 uV, v_off_fin = -89.714 uV 13:30:53 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 194.205 uVs 13:30:56 I2Y Average and RMS of Measurements VT = -172.515 +- 4.528 uVs I2Y = -15.872 +- 0.417 uTm^2 13:31:00 Wire Position: X0 = 0.0000 mm, Y0 = 0.5000 mm 13:31:04 Integrated Voltage Measurement... 13:31:13 Voltage Offset Correction: v_off_ini = -85.145 uV, v_off_fin = -89.944 uV 13:31:13 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -9.192 uVs 13:31:13 Integrated Voltage Measurement... 13:31:22 Voltage Offset Correction: v_off_ini = -77.577 uV, v_off_fin = -89.348 uV 13:31:22 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 23.264 uVs 13:31:22 Integrated Voltage Measurement... 13:31:31 Voltage Offset Correction: v_off_ini = -91.773 uV, v_off_fin = -95.367 uV 13:31:31 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -23.851 uVs 13:31:31 Integrated Voltage Measurement... 13:31:40 Voltage Offset Correction: v_off_ini = -102.644 uV, v_off_fin = -91.438 uV 13:31:40 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 17.520 uVs 13:31:40 Integrated Voltage Measurement... 13:31:49 Voltage Offset Correction: v_off_ini = -87.115 uV, v_off_fin = -106.411 uV 13:31:49 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.4999 mm dXY = -0.4999 mm VT = -38.424 uVs 13:31:49 Integrated Voltage Measurement... 13:31:58 Voltage Offset Correction: v_off_ini = -118.028 uV, v_off_fin = -103.618 uV 13:31:58 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = 29.594 uVs 13:32:02 I1X Average and RMS of Measurements VT = -23.641 +- 7.554 uVs I1X = 4.729 +- 1.511 uTm 13:32:05 Wire Position: X0 = 0.0000 mm, Y0 = 0.5000 mm 13:32:09 Integrated Voltage Measurement... 13:32:18 Voltage Offset Correction: v_off_ini = -110.058 uV, v_off_fin = -116.348 uV 13:32:18 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = -145.844 uVs 13:32:18 Integrated Voltage Measurement... 13:32:27 Voltage Offset Correction: v_off_ini = -110.458 uV, v_off_fin = -91.475 uV 13:32:27 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 157.779 uVs 13:32:27 Integrated Voltage Measurement... 13:32:36 Voltage Offset Correction: v_off_ini = -84.105 uV, v_off_fin = -117.278 uV 13:32:36 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -173.879 uVs 13:32:36 Integrated Voltage Measurement... 13:32:45 Voltage Offset Correction: v_off_ini = -107.237 uV, v_off_fin = -93.181 uV 13:32:45 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 141.595 uVs 13:32:45 Integrated Voltage Measurement... 13:32:54 Voltage Offset Correction: v_off_ini = -135.486 uV, v_off_fin = -136.257 uV 13:32:54 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.4999 mm dXY = 0.4999 mm VT = -159.263 uVs 13:32:54 Integrated Voltage Measurement... 13:33:03 Voltage Offset Correction: v_off_ini = -128.598 uV, v_off_fin = -117.459 uV 13:33:03 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 149.603 uVs 13:33:07 I1Y Average and RMS of Measurements VT = -154.660 +- 2.425 uVs I1Y = -30.934 +- 0.484 uTm 13:33:10 Wire Position: X0 = 0.0000 mm, Y0 = 0.5000 mm 13:33:12 Integrated Voltage Measurement... 13:33:21 Voltage Offset Correction: v_off_ini = -116.621 uV, v_off_fin = -108.834 uV 13:33:21 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -42.818 uVs 13:33:22 Integrated Voltage Measurement... 13:33:30 Voltage Offset Correction: v_off_ini = -124.151 uV, v_off_fin = -124.708 uV 13:33:31 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 37.805 uVs 13:33:31 Integrated Voltage Measurement... 13:33:40 Voltage Offset Correction: v_off_ini = -104.684 uV, v_off_fin = -102.912 uV 13:33:40 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -29.845 uVs 13:33:40 Integrated Voltage Measurement... 13:33:49 Voltage Offset Correction: v_off_ini = -121.159 uV, v_off_fin = -103.483 uV 13:33:49 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 35.520 uVs 13:33:49 Integrated Voltage Measurement... 13:33:58 Voltage Offset Correction: v_off_ini = -112.227 uV, v_off_fin = -91.077 uV 13:33:58 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -47.685 uVs 13:33:58 Integrated Voltage Measurement... 13:34:07 Voltage Offset Correction: v_off_ini = -105.498 uV, v_off_fin = -109.870 uV 13:34:07 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 46.747 uVs 13:34:09 I2X Average and RMS of Measurements VT = -40.070 +- 5.936 uVs I2X = 3.686 +- 0.546 uTm^2 13:34:12 Wire Position: X0 = 0.0000 mm, Y0 = 0.5000 mm 13:34:14 Integrated Voltage Measurement... 13:34:23 Voltage Offset Correction: v_off_ini = -89.371 uV, v_off_fin = -97.369 uV 13:34:23 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -172.890 uVs 13:34:23 Integrated Voltage Measurement... 13:34:32 Voltage Offset Correction: v_off_ini = -94.236 uV, v_off_fin = -104.737 uV 13:34:32 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 189.706 uVs 13:34:32 Integrated Voltage Measurement... 13:34:41 Voltage Offset Correction: v_off_ini = -88.144 uV, v_off_fin = -96.488 uV 13:34:41 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -198.443 uVs 13:34:41 Integrated Voltage Measurement... 13:34:50 Voltage Offset Correction: v_off_ini = -103.934 uV, v_off_fin = -97.124 uV 13:34:50 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 173.111 uVs 13:34:50 Integrated Voltage Measurement... 13:34:59 Voltage Offset Correction: v_off_ini = -97.195 uV, v_off_fin = -93.153 uV 13:34:59 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -176.847 uVs 13:34:59 Integrated Voltage Measurement... 13:35:08 Voltage Offset Correction: v_off_ini = -89.779 uV, v_off_fin = -98.801 uV 13:35:08 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = 167.921 uVs 13:35:11 I2Y Average and RMS of Measurements VT = -179.820 +- 5.567 uVs I2Y = -16.544 +- 0.512 uTm^2 13:35:15 Wire Position: X0 = 0.0000 mm, Y0 = 1.0000 mm 13:35:19 Integrated Voltage Measurement... 13:35:28 Voltage Offset Correction: v_off_ini = -116.582 uV, v_off_fin = -99.387 uV 13:35:28 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -31.366 uVs 13:35:28 Integrated Voltage Measurement... 13:35:37 Voltage Offset Correction: v_off_ini = -116.630 uV, v_off_fin = -103.825 uV 13:35:37 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 27.262 uVs 13:35:37 Integrated Voltage Measurement... 13:35:46 Voltage Offset Correction: v_off_ini = -100.196 uV, v_off_fin = -109.449 uV 13:35:46 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -28.093 uVs 13:35:46 Integrated Voltage Measurement... 13:35:55 Voltage Offset Correction: v_off_ini = -103.107 uV, v_off_fin = -99.401 uV 13:35:55 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 4.936 uVs 13:35:55 Integrated Voltage Measurement... 13:36:04 Voltage Offset Correction: v_off_ini = -97.821 uV, v_off_fin = -122.403 uV 13:36:04 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -44.039 uVs 13:36:04 Integrated Voltage Measurement... 13:36:13 Voltage Offset Correction: v_off_ini = -111.513 uV, v_off_fin = -117.250 uV 13:36:13 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -6.146 uVs 13:36:17 I1X Average and RMS of Measurements VT = -21.592 +- 5.550 uVs I1X = 4.318 +- 1.110 uTm 13:36:20 Wire Position: X0 = 0.0000 mm, Y0 = 1.0000 mm 13:36:24 Integrated Voltage Measurement... 13:36:33 Voltage Offset Correction: v_off_ini = -121.296 uV, v_off_fin = -132.321 uV 13:36:33 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -156.513 uVs 13:36:33 Integrated Voltage Measurement... 13:36:42 Voltage Offset Correction: v_off_ini = -113.557 uV, v_off_fin = -112.592 uV 13:36:42 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 147.357 uVs 13:36:42 Integrated Voltage Measurement... 13:36:51 Voltage Offset Correction: v_off_ini = -115.545 uV, v_off_fin = -146.578 uV 13:36:51 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -149.106 uVs 13:36:51 Integrated Voltage Measurement... 13:37:00 Voltage Offset Correction: v_off_ini = -119.255 uV, v_off_fin = -99.545 uV 13:37:00 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 163.427 uVs 13:37:00 Integrated Voltage Measurement... 13:37:09 Voltage Offset Correction: v_off_ini = -105.933 uV, v_off_fin = -126.405 uV 13:37:09 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -162.880 uVs 13:37:09 Integrated Voltage Measurement... 13:37:18 Voltage Offset Correction: v_off_ini = -130.071 uV, v_off_fin = -110.950 uV 13:37:18 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 156.413 uVs 13:37:22 I1Y Average and RMS of Measurements VT = -155.949 +- 3.156 uVs I1Y = -31.189 +- 0.631 uTm 13:37:25 Wire Position: X0 = 0.0000 mm, Y0 = 1.0000 mm 13:37:27 Integrated Voltage Measurement... 13:37:36 Voltage Offset Correction: v_off_ini = -136.469 uV, v_off_fin = -123.859 uV 13:37:36 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -26.635 uVs 13:37:37 Integrated Voltage Measurement... 13:37:45 Voltage Offset Correction: v_off_ini = -139.454 uV, v_off_fin = -134.420 uV 13:37:46 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 23.731 uVs 13:37:46 Integrated Voltage Measurement... 13:37:55 Voltage Offset Correction: v_off_ini = -123.506 uV, v_off_fin = -116.862 uV 13:37:55 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -22.924 uVs 13:37:55 Integrated Voltage Measurement... 13:38:04 Voltage Offset Correction: v_off_ini = -119.582 uV, v_off_fin = -141.525 uV 13:38:04 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 21.104 uVs 13:38:04 Integrated Voltage Measurement... 13:38:13 Voltage Offset Correction: v_off_ini = -138.964 uV, v_off_fin = -115.088 uV 13:38:13 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -35.966 uVs 13:38:13 Integrated Voltage Measurement... 13:38:22 Voltage Offset Correction: v_off_ini = -139.276 uV, v_off_fin = -153.699 uV 13:38:22 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 43.645 uVs 13:38:24 I2X Average and RMS of Measurements VT = -29.001 +- 7.749 uVs I2X = 2.668 +- 0.713 uTm^2 13:38:27 Wire Position: X0 = 0.0000 mm, Y0 = 1.0000 mm 13:38:29 Integrated Voltage Measurement... 13:38:38 Voltage Offset Correction: v_off_ini = -155.379 uV, v_off_fin = -141.987 uV 13:38:38 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -159.580 uVs 13:38:38 Integrated Voltage Measurement... 13:38:47 Voltage Offset Correction: v_off_ini = -142.675 uV, v_off_fin = -135.925 uV 13:38:47 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 203.798 uVs 13:38:47 Integrated Voltage Measurement... 13:38:56 Voltage Offset Correction: v_off_ini = -126.689 uV, v_off_fin = -138.103 uV 13:38:56 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -189.146 uVs 13:38:56 Integrated Voltage Measurement... 13:39:05 Voltage Offset Correction: v_off_ini = -129.722 uV, v_off_fin = -126.762 uV 13:39:05 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 170.900 uVs 13:39:05 Integrated Voltage Measurement... 13:39:14 Voltage Offset Correction: v_off_ini = -138.606 uV, v_off_fin = -138.187 uV 13:39:14 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -172.041 uVs 13:39:14 Integrated Voltage Measurement... 13:39:23 Voltage Offset Correction: v_off_ini = -140.872 uV, v_off_fin = -134.526 uV 13:39:23 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 165.371 uVs 13:39:26 I2Y Average and RMS of Measurements VT = -176.806 +- 5.768 uVs I2Y = -16.266 +- 0.531 uTm^2 13:39:53 Wire Position: X0 = -2.0000 mm, Y0 = 0.0000 mm 13:39:57 Integrated Voltage Measurement... 13:40:06 Voltage Offset Correction: v_off_ini = -153.981 uV, v_off_fin = -162.044 uV 13:40:06 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -59.467 uVs 13:40:06 Integrated Voltage Measurement... 13:40:15 Voltage Offset Correction: v_off_ini = -150.404 uV, v_off_fin = -141.786 uV 13:40:15 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 49.949 uVs 13:40:15 Integrated Voltage Measurement... 13:40:24 Voltage Offset Correction: v_off_ini = -148.445 uV, v_off_fin = -145.627 uV 13:40:24 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -68.655 uVs 13:40:24 Integrated Voltage Measurement... 13:40:33 Voltage Offset Correction: v_off_ini = -148.328 uV, v_off_fin = -134.127 uV 13:40:33 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 74.238 uVs 13:40:33 Integrated Voltage Measurement... 13:40:42 Voltage Offset Correction: v_off_ini = -143.256 uV, v_off_fin = -156.819 uV 13:40:42 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -60.140 uVs 13:40:42 Integrated Voltage Measurement... 13:40:51 Voltage Offset Correction: v_off_ini = -148.016 uV, v_off_fin = -142.636 uV 13:40:51 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 54.311 uVs 13:40:55 I1X Average and RMS of Measurements VT = -61.127 +- 7.369 uVs I1X = 12.225 +- 1.474 uTm 13:40:58 Wire Position: X0 = -2.0000 mm, Y0 = 0.0000 mm 13:41:02 Integrated Voltage Measurement... 13:41:11 Voltage Offset Correction: v_off_ini = -149.499 uV, v_off_fin = -145.837 uV 13:41:11 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -241.177 uVs 13:41:11 Integrated Voltage Measurement... 13:41:20 Voltage Offset Correction: v_off_ini = -136.374 uV, v_off_fin = -129.436 uV 13:41:20 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 244.821 uVs 13:41:20 Integrated Voltage Measurement... 13:41:29 Voltage Offset Correction: v_off_ini = -141.927 uV, v_off_fin = -160.975 uV 13:41:29 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -241.733 uVs 13:41:29 Integrated Voltage Measurement... 13:41:38 Voltage Offset Correction: v_off_ini = -139.854 uV, v_off_fin = -134.967 uV 13:41:38 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 251.327 uVs 13:41:38 Integrated Voltage Measurement... 13:41:47 Voltage Offset Correction: v_off_ini = -124.267 uV, v_off_fin = -139.534 uV 13:41:47 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.4999 mm dXY = 0.4999 mm VT = -260.712 uVs 13:41:47 Integrated Voltage Measurement... 13:41:56 Voltage Offset Correction: v_off_ini = -133.194 uV, v_off_fin = -120.488 uV 13:41:56 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 231.662 uVs 13:42:00 I1Y Average and RMS of Measurements VT = -245.239 +- 1.590 uVs I1Y = -49.049 +- 0.319 uTm 13:42:05 Wire Position: X0 = -2.0000 mm, Y0 = 0.0000 mm 13:42:08 Integrated Voltage Measurement... 13:42:17 Voltage Offset Correction: v_off_ini = -132.885 uV, v_off_fin = -123.578 uV 13:42:17 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -70.061 uVs 13:42:17 Integrated Voltage Measurement... 13:42:26 Voltage Offset Correction: v_off_ini = -115.671 uV, v_off_fin = -119.771 uV 13:42:26 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 76.288 uVs 13:42:26 Integrated Voltage Measurement... 13:42:35 Voltage Offset Correction: v_off_ini = -111.209 uV, v_off_fin = -100.342 uV 13:42:35 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -78.353 uVs 13:42:35 Integrated Voltage Measurement... 13:42:44 Voltage Offset Correction: v_off_ini = -102.327 uV, v_off_fin = -112.014 uV 13:42:44 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 55.478 uVs 13:42:44 Integrated Voltage Measurement... 13:42:53 Voltage Offset Correction: v_off_ini = -109.797 uV, v_off_fin = -123.674 uV 13:42:53 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -70.341 uVs 13:42:53 Integrated Voltage Measurement... 13:43:02 Voltage Offset Correction: v_off_ini = -105.985 uV, v_off_fin = -128.320 uV 13:43:02 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 51.584 uVs 13:43:05 I2X Average and RMS of Measurements VT = -67.018 +- 4.986 uVs I2X = 6.166 +- 0.459 uTm^2 13:43:10 Wire Position: X0 = -2.0000 mm, Y0 = 0.0000 mm 13:43:12 Integrated Voltage Measurement... 13:43:21 Voltage Offset Correction: v_off_ini = -126.114 uV, v_off_fin = -126.608 uV 13:43:21 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = -261.594 uVs 13:43:21 Integrated Voltage Measurement... 13:43:30 Voltage Offset Correction: v_off_ini = -115.517 uV, v_off_fin = -146.649 uV 13:43:30 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 259.755 uVs 13:43:30 Integrated Voltage Measurement... 13:43:39 Voltage Offset Correction: v_off_ini = -145.249 uV, v_off_fin = -147.361 uV 13:43:39 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = -254.214 uVs 13:43:39 Integrated Voltage Measurement... 13:43:48 Voltage Offset Correction: v_off_ini = -141.985 uV, v_off_fin = -146.790 uV 13:43:49 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 256.697 uVs 13:43:49 Integrated Voltage Measurement... 13:43:58 Voltage Offset Correction: v_off_ini = -137.784 uV, v_off_fin = -140.973 uV 13:43:58 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -251.015 uVs 13:43:58 Integrated Voltage Measurement... 13:44:07 Voltage Offset Correction: v_off_ini = -147.855 uV, v_off_fin = -131.143 uV 13:44:07 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 275.989 uVs 13:44:09 I2Y Average and RMS of Measurements VT = -259.877 +- 3.333 uVs I2Y = -23.909 +- 0.306 uTm^2 13:44:14 Wire Position: X0 = -1.0000 mm, Y0 = 0.0000 mm 13:44:18 Integrated Voltage Measurement... 13:44:27 Voltage Offset Correction: v_off_ini = -151.363 uV, v_off_fin = -143.691 uV 13:44:27 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -69.228 uVs 13:44:27 Integrated Voltage Measurement... 13:44:36 Voltage Offset Correction: v_off_ini = -131.568 uV, v_off_fin = -135.199 uV 13:44:36 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 50.340 uVs 13:44:36 Integrated Voltage Measurement... 13:44:45 Voltage Offset Correction: v_off_ini = -149.202 uV, v_off_fin = -136.873 uV 13:44:45 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -68.222 uVs 13:44:45 Integrated Voltage Measurement... 13:44:54 Voltage Offset Correction: v_off_ini = -127.021 uV, v_off_fin = -129.170 uV 13:44:54 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 43.898 uVs 13:44:54 Integrated Voltage Measurement... 13:45:03 Voltage Offset Correction: v_off_ini = -140.779 uV, v_off_fin = -143.120 uV 13:45:03 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -69.413 uVs 13:45:03 Integrated Voltage Measurement... 13:45:12 Voltage Offset Correction: v_off_ini = -148.018 uV, v_off_fin = -143.748 uV 13:45:12 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 55.207 uVs 13:45:16 I1X Average and RMS of Measurements VT = -59.385 +- 2.567 uVs I1X = 11.877 +- 0.513 uTm 13:45:19 Wire Position: X0 = -1.0000 mm, Y0 = 0.0000 mm 13:45:23 Integrated Voltage Measurement... 13:45:32 Voltage Offset Correction: v_off_ini = -146.440 uV, v_off_fin = -141.516 uV 13:45:32 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -252.490 uVs 13:45:32 Integrated Voltage Measurement... 13:45:41 Voltage Offset Correction: v_off_ini = -116.923 uV, v_off_fin = -126.787 uV 13:45:41 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 236.940 uVs 13:45:41 Integrated Voltage Measurement... 13:45:50 Voltage Offset Correction: v_off_ini = -144.868 uV, v_off_fin = -143.312 uV 13:45:50 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.4999 mm dXY = 0.4999 mm VT = -240.065 uVs 13:45:50 Integrated Voltage Measurement... 13:45:59 Voltage Offset Correction: v_off_ini = -147.073 uV, v_off_fin = -134.175 uV 13:45:59 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.4999 mm dXY = -0.4999 mm VT = 233.964 uVs 13:45:59 Integrated Voltage Measurement... 13:46:08 Voltage Offset Correction: v_off_ini = -141.435 uV, v_off_fin = -153.999 uV 13:46:08 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -225.546 uVs 13:46:08 Integrated Voltage Measurement... 13:46:17 Voltage Offset Correction: v_off_ini = -137.181 uV, v_off_fin = -147.088 uV 13:46:17 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 231.282 uVs 13:46:21 I1Y Average and RMS of Measurements VT = -236.714 +- 6.658 uVs I1Y = -47.346 +- 1.332 uTm 13:46:26 Wire Position: X0 = -1.0000 mm, Y0 = 0.0000 mm 13:46:29 Integrated Voltage Measurement... 13:46:38 Voltage Offset Correction: v_off_ini = -152.618 uV, v_off_fin = -142.469 uV 13:46:38 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -80.055 uVs 13:46:38 Integrated Voltage Measurement... 13:46:47 Voltage Offset Correction: v_off_ini = -145.816 uV, v_off_fin = -160.934 uV 13:46:47 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 60.698 uVs 13:46:47 Integrated Voltage Measurement... 13:46:56 Voltage Offset Correction: v_off_ini = -155.462 uV, v_off_fin = -157.833 uV 13:46:56 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -78.482 uVs 13:46:56 Integrated Voltage Measurement... 13:47:05 Voltage Offset Correction: v_off_ini = -158.647 uV, v_off_fin = -175.305 uV 13:47:05 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 77.672 uVs 13:47:05 Integrated Voltage Measurement... 13:47:14 Voltage Offset Correction: v_off_ini = -181.563 uV, v_off_fin = -163.712 uV 13:47:14 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -57.623 uVs 13:47:14 Integrated Voltage Measurement... 13:47:23 Voltage Offset Correction: v_off_ini = -159.348 uV, v_off_fin = -172.242 uV 13:47:23 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 56.393 uVs 13:47:26 I2X Average and RMS of Measurements VT = -68.487 +- 8.704 uVs I2X = 6.301 +- 0.801 uTm^2 13:47:31 Wire Position: X0 = -1.0000 mm, Y0 = 0.0000 mm 13:47:33 Integrated Voltage Measurement... 13:47:42 Voltage Offset Correction: v_off_ini = -161.281 uV, v_off_fin = -157.791 uV 13:47:42 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -248.442 uVs 13:47:42 Integrated Voltage Measurement... 13:47:51 Voltage Offset Correction: v_off_ini = -157.791 uV, v_off_fin = -168.522 uV 13:47:51 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 246.062 uVs 13:47:51 Integrated Voltage Measurement... 13:48:00 Voltage Offset Correction: v_off_ini = -170.290 uV, v_off_fin = -173.671 uV 13:48:00 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = -255.317 uVs 13:48:00 Integrated Voltage Measurement... 13:48:09 Voltage Offset Correction: v_off_ini = -169.687 uV, v_off_fin = -161.541 uV 13:48:09 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 240.183 uVs 13:48:09 Integrated Voltage Measurement... 13:48:18 Voltage Offset Correction: v_off_ini = -169.064 uV, v_off_fin = -160.111 uV 13:48:18 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -248.531 uVs 13:48:19 Integrated Voltage Measurement... 13:48:27 Voltage Offset Correction: v_off_ini = -156.400 uV, v_off_fin = -158.055 uV 13:48:28 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 249.286 uVs 13:48:30 I2Y Average and RMS of Measurements VT = -247.970 +- 0.694 uVs I2Y = -22.814 +- 0.064 uTm^2 13:48:35 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 13:48:39 Integrated Voltage Measurement... 13:48:48 Voltage Offset Correction: v_off_ini = -169.109 uV, v_off_fin = -170.341 uV 13:48:48 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -38.697 uVs 13:48:48 Integrated Voltage Measurement... 13:48:57 Voltage Offset Correction: v_off_ini = -177.930 uV, v_off_fin = -173.687 uV 13:48:57 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 60.121 uVs 13:48:57 Integrated Voltage Measurement... 13:49:06 Voltage Offset Correction: v_off_ini = -161.994 uV, v_off_fin = -162.320 uV 13:49:06 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -76.073 uVs 13:49:06 Integrated Voltage Measurement... 13:49:15 Voltage Offset Correction: v_off_ini = -162.940 uV, v_off_fin = -155.564 uV 13:49:15 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 50.027 uVs 13:49:15 Integrated Voltage Measurement... 13:49:24 Voltage Offset Correction: v_off_ini = -174.576 uV, v_off_fin = -171.687 uV 13:49:24 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -45.169 uVs 13:49:24 Integrated Voltage Measurement... 13:49:33 Voltage Offset Correction: v_off_ini = -170.848 uV, v_off_fin = -154.653 uV 13:49:33 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 44.868 uVs 13:49:37 I1X Average and RMS of Measurements VT = -52.493 +- 7.677 uVs I1X = 10.499 +- 1.535 uTm 13:49:40 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 13:49:44 Integrated Voltage Measurement... 13:49:53 Voltage Offset Correction: v_off_ini = -163.759 uV, v_off_fin = -175.621 uV 13:49:53 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.4999 mm dXY = 0.4999 mm VT = -256.956 uVs 13:49:53 Integrated Voltage Measurement... 13:50:02 Voltage Offset Correction: v_off_ini = -152.493 uV, v_off_fin = -153.918 uV 13:50:02 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 242.155 uVs 13:50:02 Integrated Voltage Measurement... 13:50:11 Voltage Offset Correction: v_off_ini = -158.732 uV, v_off_fin = -173.703 uV 13:50:11 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -258.186 uVs 13:50:11 Integrated Voltage Measurement... 13:50:20 Voltage Offset Correction: v_off_ini = -159.948 uV, v_off_fin = -146.281 uV 13:50:20 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 256.684 uVs 13:50:20 Integrated Voltage Measurement... 13:50:29 Voltage Offset Correction: v_off_ini = -177.799 uV, v_off_fin = -173.299 uV 13:50:29 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -234.922 uVs 13:50:29 Integrated Voltage Measurement... 13:50:38 Voltage Offset Correction: v_off_ini = -165.383 uV, v_off_fin = -149.786 uV 13:50:38 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 241.253 uVs 13:50:42 I1Y Average and RMS of Measurements VT = -248.359 +- 7.944 uVs I1Y = -49.674 +- 1.589 uTm 13:50:47 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 13:50:50 Integrated Voltage Measurement... 13:50:59 Voltage Offset Correction: v_off_ini = -181.550 uV, v_off_fin = -169.528 uV 13:50:59 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -68.941 uVs 13:50:59 Integrated Voltage Measurement... 13:51:08 Voltage Offset Correction: v_off_ini = -140.725 uV, v_off_fin = -175.705 uV 13:51:08 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 66.380 uVs 13:51:08 Integrated Voltage Measurement... 13:51:17 Voltage Offset Correction: v_off_ini = -166.472 uV, v_off_fin = -158.567 uV 13:51:17 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -65.180 uVs 13:51:17 Integrated Voltage Measurement... 13:51:26 Voltage Offset Correction: v_off_ini = -165.071 uV, v_off_fin = -170.880 uV 13:51:26 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 55.760 uVs 13:51:26 Integrated Voltage Measurement... 13:51:35 Voltage Offset Correction: v_off_ini = -168.603 uV, v_off_fin = -158.135 uV 13:51:35 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -64.802 uVs 13:51:35 Integrated Voltage Measurement... 13:51:44 Voltage Offset Correction: v_off_ini = -146.465 uV, v_off_fin = -180.273 uV 13:51:44 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 69.671 uVs 13:51:46 I2X Average and RMS of Measurements VT = -65.122 +- 3.294 uVs I2X = 5.991 +- 0.303 uTm^2 13:51:52 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 13:51:54 Integrated Voltage Measurement... 13:52:03 Voltage Offset Correction: v_off_ini = -172.933 uV, v_off_fin = -173.237 uV 13:52:03 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -261.341 uVs 13:52:03 Integrated Voltage Measurement... 13:52:12 Voltage Offset Correction: v_off_ini = -174.835 uV, v_off_fin = -177.399 uV 13:52:12 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 253.243 uVs 13:52:12 Integrated Voltage Measurement... 13:52:21 Voltage Offset Correction: v_off_ini = -161.925 uV, v_off_fin = -178.381 uV 13:52:21 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -254.690 uVs 13:52:21 Integrated Voltage Measurement... 13:52:30 Voltage Offset Correction: v_off_ini = -165.078 uV, v_off_fin = -169.644 uV 13:52:30 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 243.004 uVs 13:52:30 Integrated Voltage Measurement... 13:52:39 Voltage Offset Correction: v_off_ini = -190.707 uV, v_off_fin = -177.519 uV 13:52:39 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -247.118 uVs 13:52:39 Integrated Voltage Measurement... 13:52:48 Voltage Offset Correction: v_off_ini = -177.965 uV, v_off_fin = -169.917 uV 13:52:48 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 241.272 uVs 13:52:51 I2Y Average and RMS of Measurements VT = -250.111 +- 5.421 uVs I2Y = -23.010 +- 0.499 uTm^2 13:52:55 Wire Position: X0 = 1.0000 mm, Y0 = 0.0000 mm 13:53:00 Integrated Voltage Measurement... 13:53:09 Voltage Offset Correction: v_off_ini = -169.114 uV, v_off_fin = -164.208 uV 13:53:09 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -47.697 uVs 13:53:09 Integrated Voltage Measurement... 13:53:18 Voltage Offset Correction: v_off_ini = -157.030 uV, v_off_fin = -165.973 uV 13:53:18 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 51.132 uVs 13:53:18 Integrated Voltage Measurement... 13:53:27 Voltage Offset Correction: v_off_ini = -161.498 uV, v_off_fin = -164.734 uV 13:53:27 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -74.125 uVs 13:53:27 Integrated Voltage Measurement... 13:53:36 Voltage Offset Correction: v_off_ini = -177.190 uV, v_off_fin = -160.187 uV 13:53:36 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 55.629 uVs 13:53:36 Integrated Voltage Measurement... 13:53:45 Voltage Offset Correction: v_off_ini = -171.752 uV, v_off_fin = -172.467 uV 13:53:45 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -46.993 uVs 13:53:45 Integrated Voltage Measurement... 13:53:54 Voltage Offset Correction: v_off_ini = -157.365 uV, v_off_fin = -157.532 uV 13:53:54 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 48.230 uVs 13:53:58 I1X Average and RMS of Measurements VT = -53.968 +- 7.749 uVs I1X = 10.794 +- 1.550 uTm 13:54:00 Wire Position: X0 = 1.0000 mm, Y0 = 0.0000 mm 13:54:05 Integrated Voltage Measurement... 13:54:14 Voltage Offset Correction: v_off_ini = -155.592 uV, v_off_fin = -162.764 uV 13:54:14 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = -235.961 uVs 13:54:14 Integrated Voltage Measurement... 13:54:23 Voltage Offset Correction: v_off_ini = -153.940 uV, v_off_fin = -149.382 uV 13:54:23 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 235.870 uVs 13:54:23 Integrated Voltage Measurement... 13:54:32 Voltage Offset Correction: v_off_ini = -158.065 uV, v_off_fin = -169.001 uV 13:54:32 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -233.566 uVs 13:54:32 Integrated Voltage Measurement... 13:54:41 Voltage Offset Correction: v_off_ini = -149.608 uV, v_off_fin = -144.244 uV 13:54:41 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 245.570 uVs 13:54:41 Integrated Voltage Measurement... 13:54:50 Voltage Offset Correction: v_off_ini = -141.849 uV, v_off_fin = -155.562 uV 13:54:50 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -232.412 uVs 13:54:50 Integrated Voltage Measurement... 13:54:59 Voltage Offset Correction: v_off_ini = -133.161 uV, v_off_fin = -133.662 uV 13:54:59 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 236.299 uVs 13:55:03 I1Y Average and RMS of Measurements VT = -236.613 +- 2.184 uVs I1Y = -47.323 +- 0.437 uTm 13:55:08 Wire Position: X0 = 1.0000 mm, Y0 = 0.0000 mm 13:55:11 Integrated Voltage Measurement... 13:55:20 Voltage Offset Correction: v_off_ini = -132.708 uV, v_off_fin = -120.289 uV 13:55:20 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -92.018 uVs 13:55:20 Integrated Voltage Measurement... 13:55:29 Voltage Offset Correction: v_off_ini = -117.861 uV, v_off_fin = -120.678 uV 13:55:29 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 66.751 uVs 13:55:29 Integrated Voltage Measurement... 13:55:38 Voltage Offset Correction: v_off_ini = -146.543 uV, v_off_fin = -120.938 uV 13:55:38 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -84.179 uVs 13:55:38 Integrated Voltage Measurement... 13:55:47 Voltage Offset Correction: v_off_ini = -145.461 uV, v_off_fin = -127.915 uV 13:55:47 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 61.002 uVs 13:55:47 Integrated Voltage Measurement... 13:55:56 Voltage Offset Correction: v_off_ini = -140.948 uV, v_off_fin = -125.642 uV 13:55:56 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -80.938 uVs 13:55:56 Integrated Voltage Measurement... 13:56:05 Voltage Offset Correction: v_off_ini = -135.397 uV, v_off_fin = -151.462 uV 13:56:05 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 74.062 uVs 13:56:07 I2X Average and RMS of Measurements VT = -76.492 +- 2.864 uVs I2X = 7.037 +- 0.263 uTm^2 13:56:12 Wire Position: X0 = 1.0000 mm, Y0 = 0.0000 mm 13:56:15 Integrated Voltage Measurement... 13:56:24 Voltage Offset Correction: v_off_ini = -145.447 uV, v_off_fin = -136.533 uV 13:56:24 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -226.568 uVs 13:56:24 Integrated Voltage Measurement... 13:56:33 Voltage Offset Correction: v_off_ini = -135.080 uV, v_off_fin = -126.281 uV 13:56:33 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 254.858 uVs 13:56:33 Integrated Voltage Measurement... 13:56:42 Voltage Offset Correction: v_off_ini = -140.887 uV, v_off_fin = -147.242 uV 13:56:42 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -235.757 uVs 13:56:42 Integrated Voltage Measurement... 13:56:51 Voltage Offset Correction: v_off_ini = -151.952 uV, v_off_fin = -151.877 uV 13:56:51 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 270.390 uVs 13:56:51 Integrated Voltage Measurement... 13:57:00 Voltage Offset Correction: v_off_ini = -147.354 uV, v_off_fin = -144.674 uV 13:57:00 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -256.684 uVs 13:57:00 Integrated Voltage Measurement... 13:57:09 Voltage Offset Correction: v_off_ini = -132.153 uV, v_off_fin = -132.099 uV 13:57:09 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = 258.716 uVs 13:57:12 I2Y Average and RMS of Measurements VT = -250.495 +- 7.171 uVs I2Y = -23.046 +- 0.660 uTm^2 13:57:16 Wire Position: X0 = 2.0000 mm, Y0 = 0.0000 mm 13:57:21 Integrated Voltage Measurement... 13:57:30 Voltage Offset Correction: v_off_ini = -158.151 uV, v_off_fin = -156.576 uV 13:57:30 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -44.202 uVs 13:57:30 Integrated Voltage Measurement... 13:57:39 Voltage Offset Correction: v_off_ini = -144.684 uV, v_off_fin = -136.462 uV 13:57:39 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 53.972 uVs 13:57:39 Integrated Voltage Measurement... 13:57:48 Voltage Offset Correction: v_off_ini = -146.490 uV, v_off_fin = -162.061 uV 13:57:48 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -71.379 uVs 13:57:48 Integrated Voltage Measurement... 13:57:57 Voltage Offset Correction: v_off_ini = -143.959 uV, v_off_fin = -138.149 uV 13:57:57 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 41.552 uVs 13:57:57 Integrated Voltage Measurement... 13:58:06 Voltage Offset Correction: v_off_ini = -139.019 uV, v_off_fin = -144.549 uV 13:58:06 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -70.733 uVs 13:58:06 Integrated Voltage Measurement... 13:58:15 Voltage Offset Correction: v_off_ini = -143.405 uV, v_off_fin = -159.601 uV 13:58:15 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 35.673 uVs 13:58:19 I1X Average and RMS of Measurements VT = -52.918 +- 3.019 uVs I1X = 10.584 +- 0.604 uTm 13:58:21 Wire Position: X0 = 2.0000 mm, Y0 = 0.0000 mm 13:58:26 Integrated Voltage Measurement... 13:58:35 Voltage Offset Correction: v_off_ini = -142.460 uV, v_off_fin = -149.192 uV 13:58:35 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.4998 mm dXY = 0.4998 mm VT = -244.158 uVs 13:58:35 Integrated Voltage Measurement... 13:58:44 Voltage Offset Correction: v_off_ini = -131.234 uV, v_off_fin = -121.845 uV 13:58:44 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 250.712 uVs 13:58:44 Integrated Voltage Measurement... 13:58:53 Voltage Offset Correction: v_off_ini = -128.736 uV, v_off_fin = -149.130 uV 13:58:53 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -257.658 uVs 13:58:53 Integrated Voltage Measurement... 13:59:02 Voltage Offset Correction: v_off_ini = -149.956 uV, v_off_fin = -129.382 uV 13:59:02 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 253.068 uVs 13:59:02 Integrated Voltage Measurement... 13:59:11 Voltage Offset Correction: v_off_ini = -141.579 uV, v_off_fin = -142.305 uV 13:59:11 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -238.912 uVs 13:59:11 Integrated Voltage Measurement... 13:59:20 Voltage Offset Correction: v_off_ini = -123.796 uV, v_off_fin = -120.812 uV 13:59:20 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 241.606 uVs 13:59:24 I1Y Average and RMS of Measurements VT = -247.685 +- 6.169 uVs I1Y = -49.540 +- 1.234 uTm 13:59:29 Wire Position: X0 = 2.0000 mm, Y0 = 0.0000 mm 13:59:32 Integrated Voltage Measurement... 13:59:41 Voltage Offset Correction: v_off_ini = -127.222 uV, v_off_fin = -149.873 uV 13:59:41 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -85.998 uVs 13:59:41 Integrated Voltage Measurement... 13:59:50 Voltage Offset Correction: v_off_ini = -143.097 uV, v_off_fin = -157.235 uV 13:59:50 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 79.528 uVs 13:59:50 Integrated Voltage Measurement... 13:59:59 Voltage Offset Correction: v_off_ini = -154.073 uV, v_off_fin = -134.812 uV 13:59:59 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -78.952 uVs 13:59:59 Integrated Voltage Measurement... 14:00:08 Voltage Offset Correction: v_off_ini = -124.018 uV, v_off_fin = -146.754 uV 14:00:08 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 73.561 uVs 14:00:08 Integrated Voltage Measurement... 14:00:17 Voltage Offset Correction: v_off_ini = -148.875 uV, v_off_fin = -145.534 uV 14:00:17 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -89.010 uVs 14:00:17 Integrated Voltage Measurement... 14:00:26 Voltage Offset Correction: v_off_ini = -151.092 uV, v_off_fin = -169.267 uV 14:00:26 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 86.928 uVs 14:00:28 I2X Average and RMS of Measurements VT = -82.330 +- 4.791 uVs I2X = 7.574 +- 0.441 uTm^2 14:00:33 Wire Position: X0 = 2.0000 mm, Y0 = 0.0000 mm 14:00:36 Integrated Voltage Measurement... 14:00:45 Voltage Offset Correction: v_off_ini = -167.665 uV, v_off_fin = -153.897 uV 14:00:45 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -259.026 uVs 14:00:45 Integrated Voltage Measurement... 14:00:54 Voltage Offset Correction: v_off_ini = -150.801 uV, v_off_fin = -158.332 uV 14:00:54 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 231.521 uVs 14:00:54 Integrated Voltage Measurement... 14:01:03 Voltage Offset Correction: v_off_ini = -153.696 uV, v_off_fin = -164.427 uV 14:01:03 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -253.849 uVs 14:01:03 Integrated Voltage Measurement... 14:01:12 Voltage Offset Correction: v_off_ini = -180.348 uV, v_off_fin = -160.616 uV 14:01:12 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 264.703 uVs 14:01:12 Integrated Voltage Measurement... 14:01:21 Voltage Offset Correction: v_off_ini = -154.403 uV, v_off_fin = -146.815 uV 14:01:21 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -267.045 uVs 14:01:21 Integrated Voltage Measurement... 14:01:30 Voltage Offset Correction: v_off_ini = -140.967 uV, v_off_fin = -155.809 uV 14:01:30 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 247.723 uVs 14:01:32 I2Y Average and RMS of Measurements VT = -253.978 +- 6.203 uVs I2Y = -23.366 +- 0.571 uTm^2 14:01:40 Wire Position: X0 = 0.0000 mm, Y0 = -1.0000 mm 14:01:44 Integrated Voltage Measurement... 14:01:53 Voltage Offset Correction: v_off_ini = -151.937 uV, v_off_fin = -144.083 uV 14:01:53 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -46.370 uVs 14:01:53 Integrated Voltage Measurement... 14:02:02 Voltage Offset Correction: v_off_ini = -159.690 uV, v_off_fin = -157.672 uV 14:02:02 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 72.722 uVs 14:02:03 Integrated Voltage Measurement... 14:02:11 Voltage Offset Correction: v_off_ini = -142.887 uV, v_off_fin = -147.023 uV 14:02:12 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -64.292 uVs 14:02:12 Integrated Voltage Measurement... 14:02:20 Voltage Offset Correction: v_off_ini = -148.543 uV, v_off_fin = -151.074 uV 14:02:21 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 45.835 uVs 14:02:21 Integrated Voltage Measurement... 14:02:30 Voltage Offset Correction: v_off_ini = -163.222 uV, v_off_fin = -156.520 uV 14:02:30 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -48.650 uVs 14:02:30 Integrated Voltage Measurement... 14:02:39 Voltage Offset Correction: v_off_ini = -162.175 uV, v_off_fin = -158.057 uV 14:02:39 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 84.319 uVs 14:02:43 I1X Average and RMS of Measurements VT = -60.365 +- 4.698 uVs I1X = 12.073 +- 0.940 uTm 14:02:45 Wire Position: X0 = 0.0000 mm, Y0 = -1.0000 mm 14:02:50 Integrated Voltage Measurement... 14:02:59 Voltage Offset Correction: v_off_ini = -158.404 uV, v_off_fin = -158.641 uV 14:02:59 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -239.113 uVs 14:02:59 Integrated Voltage Measurement... 14:03:08 Voltage Offset Correction: v_off_ini = -147.187 uV, v_off_fin = -130.800 uV 14:03:08 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 236.636 uVs 14:03:08 Integrated Voltage Measurement... 14:03:17 Voltage Offset Correction: v_off_ini = -154.376 uV, v_off_fin = -162.798 uV 14:03:17 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -242.749 uVs 14:03:17 Integrated Voltage Measurement... 14:03:26 Voltage Offset Correction: v_off_ini = -163.420 uV, v_off_fin = -131.112 uV 14:03:26 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 252.744 uVs 14:03:26 Integrated Voltage Measurement... 14:03:35 Voltage Offset Correction: v_off_ini = -155.749 uV, v_off_fin = -159.030 uV 14:03:35 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -249.534 uVs 14:03:35 Integrated Voltage Measurement... 14:03:44 Voltage Offset Correction: v_off_ini = -152.403 uV, v_off_fin = -137.888 uV 14:03:44 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 248.258 uVs 14:03:48 I1Y Average and RMS of Measurements VT = -244.839 +- 4.947 uVs I1Y = -48.968 +- 0.989 uTm 14:03:50 Wire Position: X0 = 0.0000 mm, Y0 = -1.0000 mm 14:03:53 Integrated Voltage Measurement... 14:04:02 Voltage Offset Correction: v_off_ini = -145.587 uV, v_off_fin = -159.445 uV 14:04:02 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -65.454 uVs 14:04:02 Integrated Voltage Measurement... 14:04:11 Voltage Offset Correction: v_off_ini = -172.084 uV, v_off_fin = -155.186 uV 14:04:11 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 45.980 uVs 14:04:11 Integrated Voltage Measurement... 14:04:20 Voltage Offset Correction: v_off_ini = -148.289 uV, v_off_fin = -140.915 uV 14:04:20 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -70.129 uVs 14:04:20 Integrated Voltage Measurement... 14:04:29 Voltage Offset Correction: v_off_ini = -135.467 uV, v_off_fin = -150.370 uV 14:04:29 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 66.758 uVs 14:04:29 Integrated Voltage Measurement... 14:04:38 Voltage Offset Correction: v_off_ini = -145.849 uV, v_off_fin = -151.002 uV 14:04:38 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -63.867 uVs 14:04:38 Integrated Voltage Measurement... 14:04:47 Voltage Offset Correction: v_off_ini = -152.454 uV, v_off_fin = -149.707 uV 14:04:47 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 59.211 uVs 14:04:50 I2X Average and RMS of Measurements VT = -61.900 +- 5.202 uVs I2X = 5.695 +- 0.478 uTm^2 14:04:52 Wire Position: X0 = 0.0000 mm, Y0 = -1.0000 mm 14:04:55 Integrated Voltage Measurement... 14:05:04 Voltage Offset Correction: v_off_ini = -134.859 uV, v_off_fin = -134.509 uV 14:05:04 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -222.510 uVs 14:05:04 Integrated Voltage Measurement... 14:05:13 Voltage Offset Correction: v_off_ini = -145.062 uV, v_off_fin = -144.023 uV 14:05:13 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 261.695 uVs 14:05:13 Integrated Voltage Measurement... 14:05:22 Voltage Offset Correction: v_off_ini = -138.138 uV, v_off_fin = -137.996 uV 14:05:22 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -276.670 uVs 14:05:22 Integrated Voltage Measurement... 14:05:31 Voltage Offset Correction: v_off_ini = -128.967 uV, v_off_fin = -125.805 uV 14:05:31 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 247.166 uVs 14:05:31 Integrated Voltage Measurement... 14:05:40 Voltage Offset Correction: v_off_ini = -126.908 uV, v_off_fin = -120.721 uV 14:05:40 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -248.970 uVs 14:05:40 Integrated Voltage Measurement... 14:05:49 Voltage Offset Correction: v_off_ini = -146.159 uV, v_off_fin = -142.840 uV 14:05:49 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 259.810 uVs 14:05:51 I2Y Average and RMS of Measurements VT = -252.803 +- 8.167 uVs I2Y = -23.258 +- 0.751 uTm^2 14:05:55 Wire Position: X0 = 0.0000 mm, Y0 = -0.5000 mm 14:06:00 Integrated Voltage Measurement... 14:06:09 Voltage Offset Correction: v_off_ini = -142.591 uV, v_off_fin = -159.626 uV 14:06:09 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -48.115 uVs 14:06:09 Integrated Voltage Measurement... 14:06:18 Voltage Offset Correction: v_off_ini = -136.571 uV, v_off_fin = -145.409 uV 14:06:18 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 70.654 uVs 14:06:18 Integrated Voltage Measurement... 14:06:27 Voltage Offset Correction: v_off_ini = -138.473 uV, v_off_fin = -125.178 uV 14:06:27 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -64.434 uVs 14:06:27 Integrated Voltage Measurement... 14:06:36 Voltage Offset Correction: v_off_ini = -130.678 uV, v_off_fin = -146.320 uV 14:06:36 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 44.345 uVs 14:06:36 Integrated Voltage Measurement... 14:06:45 Voltage Offset Correction: v_off_ini = -141.619 uV, v_off_fin = -147.709 uV 14:06:45 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -72.583 uVs 14:06:45 Integrated Voltage Measurement... 14:06:54 Voltage Offset Correction: v_off_ini = -140.960 uV, v_off_fin = -139.772 uV 14:06:54 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 44.598 uVs 14:06:58 I1X Average and RMS of Measurements VT = -57.455 +- 2.192 uVs I1X = 11.491 +- 0.438 uTm 14:07:00 Wire Position: X0 = 0.0000 mm, Y0 = -0.5000 mm 14:07:05 Integrated Voltage Measurement... 14:07:14 Voltage Offset Correction: v_off_ini = -136.590 uV, v_off_fin = -140.654 uV 14:07:14 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -236.877 uVs 14:07:14 Integrated Voltage Measurement... 14:07:23 Voltage Offset Correction: v_off_ini = -137.551 uV, v_off_fin = -135.815 uV 14:07:23 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 232.488 uVs 14:07:23 Integrated Voltage Measurement... 14:07:32 Voltage Offset Correction: v_off_ini = -147.233 uV, v_off_fin = -152.438 uV 14:07:32 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -253.857 uVs 14:07:32 Integrated Voltage Measurement... 14:07:41 Voltage Offset Correction: v_off_ini = -139.737 uV, v_off_fin = -131.157 uV 14:07:41 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 254.094 uVs 14:07:41 Integrated Voltage Measurement... 14:07:50 Voltage Offset Correction: v_off_ini = -141.140 uV, v_off_fin = -135.131 uV 14:07:50 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -244.079 uVs 14:07:50 Integrated Voltage Measurement... 14:07:59 Voltage Offset Correction: v_off_ini = -143.157 uV, v_off_fin = -126.053 uV 14:07:59 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 242.177 uVs 14:08:03 I1Y Average and RMS of Measurements VT = -243.929 +- 7.897 uVs I1Y = -48.786 +- 1.579 uTm 14:08:05 Wire Position: X0 = 0.0000 mm, Y0 = -0.5000 mm 14:08:08 Integrated Voltage Measurement... 14:08:17 Voltage Offset Correction: v_off_ini = -118.044 uV, v_off_fin = -119.252 uV 14:08:17 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -77.870 uVs 14:08:17 Integrated Voltage Measurement... 14:08:26 Voltage Offset Correction: v_off_ini = -123.538 uV, v_off_fin = -128.734 uV 14:08:26 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 65.636 uVs 14:08:26 Integrated Voltage Measurement... 14:08:35 Voltage Offset Correction: v_off_ini = -117.437 uV, v_off_fin = -123.768 uV 14:08:35 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -79.420 uVs 14:08:35 Integrated Voltage Measurement... 14:08:44 Voltage Offset Correction: v_off_ini = -127.201 uV, v_off_fin = -136.543 uV 14:08:44 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 77.767 uVs 14:08:44 Integrated Voltage Measurement... 14:08:53 Voltage Offset Correction: v_off_ini = -118.797 uV, v_off_fin = -119.103 uV 14:08:53 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -67.314 uVs 14:08:53 Integrated Voltage Measurement... 14:09:02 Voltage Offset Correction: v_off_ini = -144.959 uV, v_off_fin = -129.216 uV 14:09:02 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 74.340 uVs 14:09:05 I2X Average and RMS of Measurements VT = -73.725 +- 3.463 uVs I2X = 6.783 +- 0.319 uTm^2 14:09:07 Wire Position: X0 = 0.0000 mm, Y0 = -0.5000 mm 14:09:10 Integrated Voltage Measurement... 14:09:19 Voltage Offset Correction: v_off_ini = -129.359 uV, v_off_fin = -121.117 uV 14:09:19 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -230.921 uVs 14:09:19 Integrated Voltage Measurement... 14:09:28 Voltage Offset Correction: v_off_ini = -120.760 uV, v_off_fin = -117.948 uV 14:09:28 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 239.960 uVs 14:09:28 Integrated Voltage Measurement... 14:09:37 Voltage Offset Correction: v_off_ini = -131.845 uV, v_off_fin = -130.423 uV 14:09:37 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -255.233 uVs 14:09:37 Integrated Voltage Measurement... 14:09:46 Voltage Offset Correction: v_off_ini = -123.903 uV, v_off_fin = -125.075 uV 14:09:46 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 230.076 uVs 14:09:46 Integrated Voltage Measurement... 14:09:55 Voltage Offset Correction: v_off_ini = -129.426 uV, v_off_fin = -128.786 uV 14:09:55 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -248.175 uVs 14:09:55 Integrated Voltage Measurement... 14:10:04 Voltage Offset Correction: v_off_ini = -123.472 uV, v_off_fin = -114.569 uV 14:10:04 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 241.519 uVs 14:10:06 I2Y Average and RMS of Measurements VT = -240.981 +- 4.018 uVs I2Y = -22.170 +- 0.370 uTm^2 14:10:10 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 14:10:15 Integrated Voltage Measurement... 14:10:23 Voltage Offset Correction: v_off_ini = -122.561 uV, v_off_fin = -119.681 uV 14:10:24 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -45.042 uVs 14:10:24 Integrated Voltage Measurement... 14:10:32 Voltage Offset Correction: v_off_ini = -139.155 uV, v_off_fin = -125.911 uV 14:10:33 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 67.034 uVs 14:10:33 Integrated Voltage Measurement... 14:10:42 Voltage Offset Correction: v_off_ini = -114.960 uV, v_off_fin = -145.603 uV 14:10:42 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -79.488 uVs 14:10:42 Integrated Voltage Measurement... 14:10:51 Voltage Offset Correction: v_off_ini = -131.774 uV, v_off_fin = -122.346 uV 14:10:51 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 65.371 uVs 14:10:51 Integrated Voltage Measurement... 14:11:00 Voltage Offset Correction: v_off_ini = -127.109 uV, v_off_fin = -123.079 uV 14:11:00 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -44.847 uVs 14:11:00 Integrated Voltage Measurement... 14:11:09 Voltage Offset Correction: v_off_ini = -125.995 uV, v_off_fin = -113.260 uV 14:11:09 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 78.417 uVs 14:11:13 I1X Average and RMS of Measurements VT = -63.366 +- 6.803 uVs I1X = 12.673 +- 1.361 uTm 14:11:15 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 14:11:20 Integrated Voltage Measurement... 14:11:28 Voltage Offset Correction: v_off_ini = -114.959 uV, v_off_fin = -133.727 uV 14:11:28 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -228.795 uVs 14:11:29 Integrated Voltage Measurement... 14:11:37 Voltage Offset Correction: v_off_ini = -132.941 uV, v_off_fin = -114.568 uV 14:11:37 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 241.848 uVs 14:11:37 Integrated Voltage Measurement... 14:11:46 Voltage Offset Correction: v_off_ini = -127.395 uV, v_off_fin = -135.572 uV 14:11:46 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -243.708 uVs 14:11:46 Integrated Voltage Measurement... 14:11:55 Voltage Offset Correction: v_off_ini = -133.510 uV, v_off_fin = -109.947 uV 14:11:55 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 236.484 uVs 14:11:55 Integrated Voltage Measurement... 14:12:04 Voltage Offset Correction: v_off_ini = -122.281 uV, v_off_fin = -137.459 uV 14:12:04 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -240.855 uVs 14:12:04 Integrated Voltage Measurement... 14:12:13 Voltage Offset Correction: v_off_ini = -140.300 uV, v_off_fin = -117.123 uV 14:12:13 Integrated Voltage Measurement dXYus = -0.5001 mm, dXYds = -0.5001 mm dXY = -0.5001 mm VT = 227.855 uVs 14:12:18 I1Y Average and RMS of Measurements VT = -236.591 +- 2.510 uVs I1Y = -47.317 +- 0.503 uTm 14:12:20 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 14:12:23 Integrated Voltage Measurement... 14:12:32 Voltage Offset Correction: v_off_ini = -117.120 uV, v_off_fin = -114.114 uV 14:12:32 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -64.423 uVs 14:12:32 Integrated Voltage Measurement... 14:12:41 Voltage Offset Correction: v_off_ini = -118.606 uV, v_off_fin = -137.040 uV 14:12:41 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 92.309 uVs 14:12:41 Integrated Voltage Measurement... 14:12:50 Voltage Offset Correction: v_off_ini = -143.406 uV, v_off_fin = -141.932 uV 14:12:50 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -54.524 uVs 14:12:50 Integrated Voltage Measurement... 14:12:59 Voltage Offset Correction: v_off_ini = -144.923 uV, v_off_fin = -144.726 uV 14:12:59 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 79.736 uVs 14:12:59 Integrated Voltage Measurement... 14:13:08 Voltage Offset Correction: v_off_ini = -141.374 uV, v_off_fin = -143.743 uV 14:13:08 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -39.370 uVs 14:13:08 Integrated Voltage Measurement... 14:13:17 Voltage Offset Correction: v_off_ini = -163.213 uV, v_off_fin = -152.551 uV 14:13:17 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 59.886 uVs 14:13:19 I2X Average and RMS of Measurements VT = -65.041 +- 11.825 uVs I2X = 5.984 +- 1.088 uTm^2 14:13:22 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 14:13:24 Integrated Voltage Measurement... 14:13:33 Voltage Offset Correction: v_off_ini = -130.615 uV, v_off_fin = -132.665 uV 14:13:33 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -246.623 uVs 14:13:33 Integrated Voltage Measurement... 14:13:42 Voltage Offset Correction: v_off_ini = -130.402 uV, v_off_fin = -149.228 uV 14:13:42 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 252.628 uVs 14:13:42 Integrated Voltage Measurement... 14:13:51 Voltage Offset Correction: v_off_ini = -132.542 uV, v_off_fin = -145.788 uV 14:13:51 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -250.301 uVs 14:13:51 Integrated Voltage Measurement... 14:14:00 Voltage Offset Correction: v_off_ini = -157.611 uV, v_off_fin = -153.304 uV 14:14:01 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 274.493 uVs 14:14:01 Integrated Voltage Measurement... 14:14:10 Voltage Offset Correction: v_off_ini = -144.115 uV, v_off_fin = -145.071 uV 14:14:10 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -246.049 uVs 14:14:10 Integrated Voltage Measurement... 14:14:19 Voltage Offset Correction: v_off_ini = -140.290 uV, v_off_fin = -151.269 uV 14:14:19 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 229.151 uVs 14:14:21 I2Y Average and RMS of Measurements VT = -249.874 +- 10.125 uVs I2Y = -22.988 +- 0.931 uTm^2 14:14:25 Wire Position: X0 = 0.0000 mm, Y0 = 0.5000 mm 14:14:29 Integrated Voltage Measurement... 14:14:38 Voltage Offset Correction: v_off_ini = -158.523 uV, v_off_fin = -170.768 uV 14:14:38 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -30.763 uVs 14:14:38 Integrated Voltage Measurement... 14:14:47 Voltage Offset Correction: v_off_ini = -154.306 uV, v_off_fin = -147.319 uV 14:14:47 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 46.615 uVs 14:14:47 Integrated Voltage Measurement... 14:14:56 Voltage Offset Correction: v_off_ini = -161.535 uV, v_off_fin = -166.116 uV 14:14:56 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -41.313 uVs 14:14:56 Integrated Voltage Measurement... 14:15:05 Voltage Offset Correction: v_off_ini = -157.860 uV, v_off_fin = -145.132 uV 14:15:05 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 63.986 uVs 14:15:05 Integrated Voltage Measurement... 14:15:14 Voltage Offset Correction: v_off_ini = -151.538 uV, v_off_fin = -140.215 uV 14:15:14 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -35.674 uVs 14:15:14 Integrated Voltage Measurement... 14:15:23 Voltage Offset Correction: v_off_ini = -146.966 uV, v_off_fin = -129.726 uV 14:15:23 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 67.828 uVs 14:15:28 I1X Average and RMS of Measurements VT = -47.696 +- 6.380 uVs I1X = 9.539 +- 1.276 uTm 14:15:30 Wire Position: X0 = 0.0000 mm, Y0 = 0.5000 mm 14:15:34 Integrated Voltage Measurement... 14:15:43 Voltage Offset Correction: v_off_ini = -142.584 uV, v_off_fin = -149.292 uV 14:15:43 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -244.050 uVs 14:15:43 Integrated Voltage Measurement... 14:15:52 Voltage Offset Correction: v_off_ini = -143.245 uV, v_off_fin = -130.633 uV 14:15:52 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 251.466 uVs 14:15:52 Integrated Voltage Measurement... 14:16:01 Voltage Offset Correction: v_off_ini = -123.122 uV, v_off_fin = -144.326 uV 14:16:01 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -249.514 uVs 14:16:01 Integrated Voltage Measurement... 14:16:10 Voltage Offset Correction: v_off_ini = -140.592 uV, v_off_fin = -138.275 uV 14:16:10 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 239.094 uVs 14:16:10 Integrated Voltage Measurement... 14:16:19 Voltage Offset Correction: v_off_ini = -147.518 uV, v_off_fin = -172.662 uV 14:16:19 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -235.024 uVs 14:16:19 Integrated Voltage Measurement... 14:16:28 Voltage Offset Correction: v_off_ini = -166.397 uV, v_off_fin = -147.144 uV 14:16:28 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 241.335 uVs 14:16:32 I1Y Average and RMS of Measurements VT = -243.414 +- 3.961 uVs I1Y = -48.682 +- 0.792 uTm 14:16:35 Wire Position: X0 = 0.0000 mm, Y0 = 0.5000 mm 14:16:37 Integrated Voltage Measurement... 14:16:46 Voltage Offset Correction: v_off_ini = -158.005 uV, v_off_fin = -149.971 uV 14:16:46 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -81.368 uVs 14:16:46 Integrated Voltage Measurement... 14:16:55 Voltage Offset Correction: v_off_ini = -148.837 uV, v_off_fin = -161.568 uV 14:16:55 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 95.504 uVs 14:16:55 Integrated Voltage Measurement... 14:17:04 Voltage Offset Correction: v_off_ini = -155.649 uV, v_off_fin = -173.734 uV 14:17:04 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -66.688 uVs 14:17:04 Integrated Voltage Measurement... 14:17:13 Voltage Offset Correction: v_off_ini = -160.742 uV, v_off_fin = -162.611 uV 14:17:13 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 78.073 uVs 14:17:13 Integrated Voltage Measurement... 14:17:22 Voltage Offset Correction: v_off_ini = -147.461 uV, v_off_fin = -142.624 uV 14:17:22 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -75.086 uVs 14:17:22 Integrated Voltage Measurement... 14:17:31 Voltage Offset Correction: v_off_ini = -148.289 uV, v_off_fin = -152.562 uV 14:17:31 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 53.362 uVs 14:17:34 I2X Average and RMS of Measurements VT = -75.014 +- 10.058 uVs I2X = 6.901 +- 0.925 uTm^2 14:17:36 Wire Position: X0 = 0.0000 mm, Y0 = 0.5000 mm 14:17:39 Integrated Voltage Measurement... 14:17:48 Voltage Offset Correction: v_off_ini = -150.492 uV, v_off_fin = -153.485 uV 14:17:48 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -248.972 uVs 14:17:48 Integrated Voltage Measurement... 14:17:57 Voltage Offset Correction: v_off_ini = -137.901 uV, v_off_fin = -144.951 uV 14:17:57 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 252.341 uVs 14:17:57 Integrated Voltage Measurement... 14:18:06 Voltage Offset Correction: v_off_ini = -149.090 uV, v_off_fin = -152.582 uV 14:18:06 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -256.463 uVs 14:18:06 Integrated Voltage Measurement... 14:18:15 Voltage Offset Correction: v_off_ini = -141.751 uV, v_off_fin = -140.585 uV 14:18:15 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 258.665 uVs 14:18:15 Integrated Voltage Measurement... 14:18:24 Voltage Offset Correction: v_off_ini = -139.892 uV, v_off_fin = -142.555 uV 14:18:24 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -247.862 uVs 14:18:24 Integrated Voltage Measurement... 14:18:33 Voltage Offset Correction: v_off_ini = -136.267 uV, v_off_fin = -131.705 uV 14:18:33 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 248.414 uVs 14:18:35 I2Y Average and RMS of Measurements VT = -252.120 +- 3.985 uVs I2Y = -23.195 +- 0.367 uTm^2 14:18:39 Wire Position: X0 = 0.0000 mm, Y0 = 1.0000 mm 14:18:44 Integrated Voltage Measurement... 14:18:53 Voltage Offset Correction: v_off_ini = -158.076 uV, v_off_fin = -139.330 uV 14:18:53 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -35.108 uVs 14:18:53 Integrated Voltage Measurement... 14:19:02 Voltage Offset Correction: v_off_ini = -139.394 uV, v_off_fin = -132.718 uV 14:19:02 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 51.329 uVs 14:19:02 Integrated Voltage Measurement... 14:19:11 Voltage Offset Correction: v_off_ini = -135.546 uV, v_off_fin = -130.999 uV 14:19:11 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -54.172 uVs 14:19:11 Integrated Voltage Measurement... 14:19:20 Voltage Offset Correction: v_off_ini = -128.799 uV, v_off_fin = -121.146 uV 14:19:20 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 62.847 uVs 14:19:20 Integrated Voltage Measurement... 14:19:29 Voltage Offset Correction: v_off_ini = -120.584 uV, v_off_fin = -143.813 uV 14:19:29 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -50.118 uVs 14:19:29 Integrated Voltage Measurement... 14:19:38 Voltage Offset Correction: v_off_ini = -142.130 uV, v_off_fin = -128.653 uV 14:19:38 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 44.997 uVs 14:19:42 I1X Average and RMS of Measurements VT = -49.762 +- 6.434 uVs I1X = 9.952 +- 1.287 uTm 14:19:44 Wire Position: X0 = 0.0000 mm, Y0 = 1.0000 mm 14:19:49 Integrated Voltage Measurement... 14:19:58 Voltage Offset Correction: v_off_ini = -127.714 uV, v_off_fin = -139.797 uV 14:19:58 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -246.547 uVs 14:19:58 Integrated Voltage Measurement... 14:20:07 Voltage Offset Correction: v_off_ini = -144.942 uV, v_off_fin = -126.463 uV 14:20:07 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 248.563 uVs 14:20:07 Integrated Voltage Measurement... 14:20:16 Voltage Offset Correction: v_off_ini = -127.507 uV, v_off_fin = -155.966 uV 14:20:16 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -241.046 uVs 14:20:16 Integrated Voltage Measurement... 14:20:24 Voltage Offset Correction: v_off_ini = -136.424 uV, v_off_fin = -127.257 uV 14:20:25 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 240.739 uVs 14:20:25 Integrated Voltage Measurement... 14:20:33 Voltage Offset Correction: v_off_ini = -140.864 uV, v_off_fin = -143.689 uV 14:20:34 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -245.301 uVs 14:20:34 Integrated Voltage Measurement... 14:20:42 Voltage Offset Correction: v_off_ini = -120.868 uV, v_off_fin = -118.257 uV 14:20:42 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 244.674 uVs 14:20:47 I1Y Average and RMS of Measurements VT = -244.478 +- 2.744 uVs I1Y = -48.896 +- 0.549 uTm 14:20:49 Wire Position: X0 = 0.0000 mm, Y0 = 1.0000 mm 14:20:52 Integrated Voltage Measurement... 14:21:01 Voltage Offset Correction: v_off_ini = -123.960 uV, v_off_fin = -120.162 uV 14:21:01 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -89.342 uVs 14:21:01 Integrated Voltage Measurement... 14:21:10 Voltage Offset Correction: v_off_ini = -124.910 uV, v_off_fin = -136.001 uV 14:21:10 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 81.169 uVs 14:21:10 Integrated Voltage Measurement... 14:21:19 Voltage Offset Correction: v_off_ini = -145.542 uV, v_off_fin = -121.855 uV 14:21:19 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -64.630 uVs 14:21:19 Integrated Voltage Measurement... 14:21:28 Voltage Offset Correction: v_off_ini = -130.628 uV, v_off_fin = -129.042 uV 14:21:28 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 66.632 uVs 14:21:28 Integrated Voltage Measurement... 14:21:37 Voltage Offset Correction: v_off_ini = -118.671 uV, v_off_fin = -115.121 uV 14:21:37 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -57.742 uVs 14:21:37 Integrated Voltage Measurement... 14:21:46 Voltage Offset Correction: v_off_ini = -117.361 uV, v_off_fin = -130.038 uV 14:21:46 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 82.331 uVs 14:21:48 I2X Average and RMS of Measurements VT = -73.641 +- 8.407 uVs I2X = 6.775 +- 0.773 uTm^2 14:21:51 Wire Position: X0 = 0.0000 mm, Y0 = 1.0000 mm 14:21:53 Integrated Voltage Measurement... 14:22:02 Voltage Offset Correction: v_off_ini = -106.023 uV, v_off_fin = -118.398 uV 14:22:02 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -251.709 uVs 14:22:02 Integrated Voltage Measurement... 14:22:11 Voltage Offset Correction: v_off_ini = -120.601 uV, v_off_fin = -104.912 uV 14:22:11 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 258.164 uVs 14:22:11 Integrated Voltage Measurement... 14:22:20 Voltage Offset Correction: v_off_ini = -123.030 uV, v_off_fin = -119.856 uV 14:22:20 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -254.706 uVs 14:22:20 Integrated Voltage Measurement... 14:22:29 Voltage Offset Correction: v_off_ini = -109.328 uV, v_off_fin = -127.472 uV 14:22:29 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 265.961 uVs 14:22:29 Integrated Voltage Measurement... 14:22:38 Voltage Offset Correction: v_off_ini = -126.680 uV, v_off_fin = -111.747 uV 14:22:38 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -255.232 uVs 14:22:38 Integrated Voltage Measurement... 14:22:47 Voltage Offset Correction: v_off_ini = -126.941 uV, v_off_fin = -109.164 uV 14:22:47 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 259.355 uVs 14:22:50 I2Y Average and RMS of Measurements VT = -257.521 +- 2.209 uVs I2Y = -23.692 +- 0.203 uTm^2 14:22:54 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 14:52:32 Wire Position: X0 = 50.0000 mm, Y0 = 0.0000 mm