SLAC Magnetic Measurements Date: 08-08-2025 Time: 11:10:51 Project: LCLS-II-HE Magnet Type: Phase Shifter Magnet Name: HE_SXPS_16342 Dataset Number: 1 Run Purpose: Tuning Run Type: Run Number: 3 Measurement Device: Stretched Wire Operator: YL Comment: 11:10:59 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:11:02 Motion Parameters, X Controller, Axis 1 acc = 0.001000 m/s/s dcl = 0.001000 m/s/s vel = 0.001000 m/s pos = 0.000000 m kp = 188.600000 ki = 820.000000 kd = 491.400000 db = 0.000000 11:11:03 Motion Parameters, X Controller, Axis 2 acc = 0.001000 m/s/s dcl = 0.001000 m/s/s vel = 0.001000 m/s pos = 0.000000 m kp = 188.600000 ki = 820.000000 kd = 491.400000 db = 0.000000 11:11:05 Motion Parameters, Y Controller, Axis 1 acc = 0.001000 m/s/s dcl = 0.001000 m/s/s vel = 0.001000 m/s pos = 0.000000 m kp = 188.600000 ki = 820.000000 kd = 491.400000 db = 0.000000 11:11:06 Motion Parameters, Y Controller, Axis 2 acc = 0.001000 m/s/s dcl = 0.001000 m/s/s vel = 0.001000 m/s pos = 0.000000 m kp = 188.600000 ki = 820.000000 kd = 491.400000 db = 0.000000 11:11:18 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:11:22 Integrated Voltage Measurement... 11:11:31 Voltage Offset Correction: v_off_ini = 375.056 uV, v_off_fin = 369.727 uV 11:11:31 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -67.219 uVs 11:11:31 Integrated Voltage Measurement... 11:11:40 Voltage Offset Correction: v_off_ini = 373.886 uV, v_off_fin = 369.770 uV 11:11:40 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 64.536 uVs 11:11:40 Integrated Voltage Measurement... 11:11:49 Voltage Offset Correction: v_off_ini = 369.014 uV, v_off_fin = 373.226 uV 11:11:49 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -82.751 uVs 11:11:49 Integrated Voltage Measurement... 11:11:58 Voltage Offset Correction: v_off_ini = 376.388 uV, v_off_fin = 378.033 uV 11:11:58 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 60.168 uVs 11:11:58 Integrated Voltage Measurement... 11:12:07 Voltage Offset Correction: v_off_ini = 381.555 uV, v_off_fin = 381.320 uV 11:12:07 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -60.154 uVs 11:12:07 Integrated Voltage Measurement... 11:12:16 Voltage Offset Correction: v_off_ini = 373.491 uV, v_off_fin = 370.653 uV 11:12:16 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 64.508 uVs 11:12:20 I1X Average and RMS of Measurements VT = -66.556 +- 3.758 uVs I1X = 13.311 +- 0.752 uTm 11:12:23 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:12:27 Integrated Voltage Measurement... 11:12:36 Voltage Offset Correction: v_off_ini = 374.969 uV, v_off_fin = 358.907 uV 11:12:36 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -254.887 uVs 11:12:36 Integrated Voltage Measurement... 11:12:45 Voltage Offset Correction: v_off_ini = 383.793 uV, v_off_fin = 384.384 uV 11:12:45 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 254.364 uVs 11:12:45 Integrated Voltage Measurement... 11:12:54 Voltage Offset Correction: v_off_ini = 365.936 uV, v_off_fin = 355.556 uV 11:12:54 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -261.649 uVs 11:12:54 Integrated Voltage Measurement... 11:13:03 Voltage Offset Correction: v_off_ini = 373.246 uV, v_off_fin = 387.627 uV 11:13:03 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 269.009 uVs 11:13:03 Integrated Voltage Measurement... 11:13:12 Voltage Offset Correction: v_off_ini = 377.031 uV, v_off_fin = 368.533 uV 11:13:12 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -276.965 uVs 11:13:12 Integrated Voltage Measurement... 11:13:21 Voltage Offset Correction: v_off_ini = 382.211 uV, v_off_fin = 396.367 uV 11:13:21 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 269.580 uVs 11:13:25 I1Y Average and RMS of Measurements VT = -264.409 +- 7.641 uVs I1Y = -52.877 +- 1.529 uTm 11:13:30 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:13:33 Integrated Voltage Measurement... 11:13:42 Voltage Offset Correction: v_off_ini = 364.923 uV, v_off_fin = 379.555 uV 11:13:42 Integrated Voltage Measurement dXYus = -1.0001 mm, dXYds = 0.0000 mm dXY = -1.0001 mm VT = -62.761 uVs 11:13:42 Integrated Voltage Measurement... 11:13:51 Voltage Offset Correction: v_off_ini = 380.216 uV, v_off_fin = 391.049 uV 11:13:51 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 44.675 uVs 11:13:51 Integrated Voltage Measurement... 11:14:00 Voltage Offset Correction: v_off_ini = 377.249 uV, v_off_fin = 365.392 uV 11:14:00 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -39.838 uVs 11:14:00 Integrated Voltage Measurement... 11:14:09 Voltage Offset Correction: v_off_ini = 364.516 uV, v_off_fin = 376.055 uV 11:14:09 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 40.467 uVs 11:14:09 Integrated Voltage Measurement... 11:14:18 Voltage Offset Correction: v_off_ini = 369.176 uV, v_off_fin = 351.944 uV 11:14:18 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -28.482 uVs 11:14:18 Integrated Voltage Measurement... 11:14:27 Voltage Offset Correction: v_off_ini = 375.676 uV, v_off_fin = 374.342 uV 11:14:27 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 33.173 uVs 11:14:29 I2X Average and RMS of Measurements VT = -41.566 +- 9.398 uVs I2X = 3.824 +- 0.864 uTm^2 11:14:34 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:14:37 Integrated Voltage Measurement... 11:14:46 Voltage Offset Correction: v_off_ini = 364.009 uV, v_off_fin = 361.050 uV 11:14:46 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -494.831 uVs 11:14:46 Integrated Voltage Measurement... 11:14:55 Voltage Offset Correction: v_off_ini = 378.152 uV, v_off_fin = 357.976 uV 11:14:55 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 480.915 uVs 11:14:55 Integrated Voltage Measurement... 11:15:04 Voltage Offset Correction: v_off_ini = 367.877 uV, v_off_fin = 355.317 uV 11:15:04 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -471.721 uVs 11:15:04 Integrated Voltage Measurement... 11:15:13 Voltage Offset Correction: v_off_ini = 364.445 uV, v_off_fin = 372.879 uV 11:15:13 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 490.703 uVs 11:15:13 Integrated Voltage Measurement... 11:15:22 Voltage Offset Correction: v_off_ini = 368.484 uV, v_off_fin = 359.541 uV 11:15:22 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -483.297 uVs 11:15:22 Integrated Voltage Measurement... 11:15:31 Voltage Offset Correction: v_off_ini = 370.764 uV, v_off_fin = 373.281 uV 11:15:31 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 490.389 uVs 11:15:34 I2Y Average and RMS of Measurements VT = -485.309 +- 2.928 uVs I2Y = -44.648 +- 0.269 uTm^2 11:15:44 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:15:49 Integrated Voltage Measurement... 11:15:57 Voltage Offset Correction: v_off_ini = 380.556 uV, v_off_fin = 379.146 uV 11:15:58 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -40.657 uVs 11:15:58 Integrated Voltage Measurement... 11:16:07 Voltage Offset Correction: v_off_ini = 375.572 uV, v_off_fin = 381.699 uV 11:16:07 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 58.726 uVs 11:16:07 Integrated Voltage Measurement... 11:16:16 Voltage Offset Correction: v_off_ini = 365.637 uV, v_off_fin = 375.154 uV 11:16:16 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -51.277 uVs 11:16:16 Integrated Voltage Measurement... 11:16:25 Voltage Offset Correction: v_off_ini = 374.688 uV, v_off_fin = 385.056 uV 11:16:25 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 76.330 uVs 11:16:25 Integrated Voltage Measurement... 11:16:34 Voltage Offset Correction: v_off_ini = 373.493 uV, v_off_fin = 367.681 uV 11:16:34 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -61.318 uVs 11:16:34 Integrated Voltage Measurement... 11:16:43 Voltage Offset Correction: v_off_ini = 390.819 uV, v_off_fin = 378.569 uV 11:16:43 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 92.127 uVs 11:16:47 I1X Average and RMS of Measurements VT = -63.406 +- 11.039 uVs I1X = 12.681 +- 2.208 uTm 11:16:49 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:16:54 Integrated Voltage Measurement... 11:17:03 Voltage Offset Correction: v_off_ini = 389.184 uV, v_off_fin = 368.183 uV 11:17:03 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = -226.852 uVs 11:17:03 Integrated Voltage Measurement... 11:17:12 Voltage Offset Correction: v_off_ini = 363.629 uV, v_off_fin = 392.953 uV 11:17:12 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 254.990 uVs 11:17:12 Integrated Voltage Measurement... 11:17:21 Voltage Offset Correction: v_off_ini = 377.405 uV, v_off_fin = 364.430 uV 11:17:21 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -253.353 uVs 11:17:21 Integrated Voltage Measurement... 11:17:30 Voltage Offset Correction: v_off_ini = 371.934 uV, v_off_fin = 392.625 uV 11:17:30 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 256.382 uVs 11:17:30 Integrated Voltage Measurement... 11:17:39 Voltage Offset Correction: v_off_ini = 376.372 uV, v_off_fin = 362.086 uV 11:17:39 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -245.099 uVs 11:17:39 Integrated Voltage Measurement... 11:17:48 Voltage Offset Correction: v_off_ini = 377.160 uV, v_off_fin = 392.966 uV 11:17:48 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 261.198 uVs 11:17:52 I1Y Average and RMS of Measurements VT = -249.645 +- 6.209 uVs I1Y = -49.926 +- 1.240 uTm 11:17:57 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:18:00 Integrated Voltage Measurement... 11:18:09 Voltage Offset Correction: v_off_ini = 386.520 uV, v_off_fin = 388.131 uV 11:18:09 Integrated Voltage Measurement dXYus = -1.0002 mm, dXYds = 0.0000 mm dXY = -1.0002 mm VT = -46.267 uVs 11:18:09 Integrated Voltage Measurement... 11:18:18 Voltage Offset Correction: v_off_ini = 387.084 uV, v_off_fin = 391.081 uV 11:18:18 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 17.774 uVs 11:18:18 Integrated Voltage Measurement... 11:18:27 Voltage Offset Correction: v_off_ini = 380.465 uV, v_off_fin = 355.029 uV 11:18:27 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -41.305 uVs 11:18:27 Integrated Voltage Measurement... 11:18:36 Voltage Offset Correction: v_off_ini = 379.235 uV, v_off_fin = 391.815 uV 11:18:36 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 51.759 uVs 11:18:36 Integrated Voltage Measurement... 11:18:45 Voltage Offset Correction: v_off_ini = 380.354 uV, v_off_fin = 377.699 uV 11:18:45 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -45.153 uVs 11:18:45 Integrated Voltage Measurement... 11:18:54 Voltage Offset Correction: v_off_ini = 388.276 uV, v_off_fin = 394.690 uV 11:18:54 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 39.395 uVs 11:18:56 I2X Average and RMS of Measurements VT = -40.275 +- 6.090 uVs I2X = 3.705 +- 0.561 uTm^2 11:19:01 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:19:04 Integrated Voltage Measurement... 11:19:13 Voltage Offset Correction: v_off_ini = 382.127 uV, v_off_fin = 388.593 uV 11:19:13 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -445.700 uVs 11:19:13 Integrated Voltage Measurement... 11:19:22 Voltage Offset Correction: v_off_ini = 394.682 uV, v_off_fin = 402.511 uV 11:19:22 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 455.817 uVs 11:19:22 Integrated Voltage Measurement... 11:19:31 Voltage Offset Correction: v_off_ini = 388.549 uV, v_off_fin = 381.598 uV 11:19:31 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -473.629 uVs 11:19:31 Integrated Voltage Measurement... 11:19:40 Voltage Offset Correction: v_off_ini = 390.835 uV, v_off_fin = 380.259 uV 11:19:40 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 434.846 uVs 11:19:40 Integrated Voltage Measurement... 11:19:49 Voltage Offset Correction: v_off_ini = 384.986 uV, v_off_fin = 386.942 uV 11:19:49 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -458.768 uVs 11:19:49 Integrated Voltage Measurement... 11:19:58 Voltage Offset Correction: v_off_ini = 383.305 uV, v_off_fin = 388.440 uV 11:19:58 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 462.483 uVs 11:20:01 I2Y Average and RMS of Measurements VT = -455.207 +- 4.086 uVs I2Y = -41.879 +- 0.376 uTm^2 11:20:11 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:20:16 Integrated Voltage Measurement... 11:20:25 Voltage Offset Correction: v_off_ini = 402.879 uV, v_off_fin = 373.890 uV 11:20:25 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -85.062 uVs 11:20:25 Integrated Voltage Measurement... 11:20:34 Voltage Offset Correction: v_off_ini = 379.801 uV, v_off_fin = 379.646 uV 11:20:34 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = 78.660 uVs 11:20:34 Integrated Voltage Measurement... 11:20:43 Voltage Offset Correction: v_off_ini = 376.934 uV, v_off_fin = 369.991 uV 11:20:43 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -80.431 uVs 11:20:43 Integrated Voltage Measurement... 11:20:52 Voltage Offset Correction: v_off_ini = 384.922 uV, v_off_fin = 399.889 uV 11:20:52 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 77.529 uVs 11:20:52 Integrated Voltage Measurement... 11:21:01 Voltage Offset Correction: v_off_ini = 397.480 uV, v_off_fin = 389.067 uV 11:21:01 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -77.896 uVs 11:21:01 Integrated Voltage Measurement... 11:21:10 Voltage Offset Correction: v_off_ini = 395.603 uV, v_off_fin = 398.869 uV 11:21:10 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 68.652 uVs 11:21:14 I1X Average and RMS of Measurements VT = -78.038 +- 3.568 uVs I1X = 15.608 +- 0.714 uTm 11:21:16 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:21:21 Integrated Voltage Measurement... 11:21:30 Voltage Offset Correction: v_off_ini = 383.115 uV, v_off_fin = 376.414 uV 11:21:30 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -224.921 uVs 11:21:30 Integrated Voltage Measurement... 11:21:39 Voltage Offset Correction: v_off_ini = 391.015 uV, v_off_fin = 409.896 uV 11:21:39 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5002 mm dXY = -0.5001 mm VT = 221.254 uVs 11:21:39 Integrated Voltage Measurement... 11:21:48 Voltage Offset Correction: v_off_ini = 395.932 uV, v_off_fin = 384.598 uV 11:21:48 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5002 mm dXY = 0.5001 mm VT = -220.200 uVs 11:21:48 Integrated Voltage Measurement... 11:21:57 Voltage Offset Correction: v_off_ini = 387.836 uV, v_off_fin = 407.325 uV 11:21:57 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 248.457 uVs 11:21:57 Integrated Voltage Measurement... 11:22:06 Voltage Offset Correction: v_off_ini = 392.926 uV, v_off_fin = 386.514 uV 11:22:06 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -226.037 uVs 11:22:06 Integrated Voltage Measurement... 11:22:15 Voltage Offset Correction: v_off_ini = 390.651 uV, v_off_fin = 413.265 uV 11:22:15 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 228.552 uVs 11:22:19 I1Y Average and RMS of Measurements VT = -228.237 +- 4.637 uVs I1Y = -45.643 +- 0.927 uTm 11:22:24 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:22:27 Integrated Voltage Measurement... 11:22:36 Voltage Offset Correction: v_off_ini = 396.850 uV, v_off_fin = 393.659 uV 11:22:36 Integrated Voltage Measurement dXYus = -1.0002 mm, dXYds = 0.0000 mm dXY = -1.0002 mm VT = -74.032 uVs 11:22:36 Integrated Voltage Measurement... 11:22:45 Voltage Offset Correction: v_off_ini = 393.096 uV, v_off_fin = 401.594 uV 11:22:45 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 42.859 uVs 11:22:45 Integrated Voltage Measurement... 11:22:54 Voltage Offset Correction: v_off_ini = 396.924 uV, v_off_fin = 399.807 uV 11:22:54 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -47.739 uVs 11:22:54 Integrated Voltage Measurement... 11:23:03 Voltage Offset Correction: v_off_ini = 417.510 uV, v_off_fin = 419.013 uV 11:23:03 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 43.130 uVs 11:23:03 Integrated Voltage Measurement... 11:23:12 Voltage Offset Correction: v_off_ini = 410.713 uV, v_off_fin = 408.965 uV 11:23:12 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -39.614 uVs 11:23:12 Integrated Voltage Measurement... 11:23:21 Voltage Offset Correction: v_off_ini = 407.943 uV, v_off_fin = 423.079 uV 11:23:21 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 59.867 uVs 11:23:23 I2X Average and RMS of Measurements VT = -51.207 +- 5.412 uVs I2X = 4.711 +- 0.497 uTm^2 11:23:28 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:23:31 Integrated Voltage Measurement... 11:23:40 Voltage Offset Correction: v_off_ini = 421.294 uV, v_off_fin = 415.024 uV 11:23:40 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -441.799 uVs 11:23:40 Integrated Voltage Measurement... 11:23:49 Voltage Offset Correction: v_off_ini = 411.669 uV, v_off_fin = 411.220 uV 11:23:49 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 413.100 uVs 11:23:49 Integrated Voltage Measurement... 11:23:58 Voltage Offset Correction: v_off_ini = 396.060 uV, v_off_fin = 383.834 uV 11:23:58 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -436.268 uVs 11:23:58 Integrated Voltage Measurement... 11:24:07 Voltage Offset Correction: v_off_ini = 384.796 uV, v_off_fin = 403.879 uV 11:24:07 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 428.900 uVs 11:24:07 Integrated Voltage Measurement... 11:24:16 Voltage Offset Correction: v_off_ini = 416.069 uV, v_off_fin = 409.770 uV 11:24:16 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -422.370 uVs 11:24:16 Integrated Voltage Measurement... 11:24:25 Voltage Offset Correction: v_off_ini = 403.118 uV, v_off_fin = 406.278 uV 11:24:25 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 433.617 uVs 11:24:28 I2Y Average and RMS of Measurements VT = -429.342 +- 2.303 uVs I2Y = -39.499 +- 0.212 uTm^2 11:24:38 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:24:43 Integrated Voltage Measurement... 11:24:52 Voltage Offset Correction: v_off_ini = 413.559 uV, v_off_fin = 402.373 uV 11:24:52 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -84.966 uVs 11:24:52 Integrated Voltage Measurement... 11:25:01 Voltage Offset Correction: v_off_ini = 420.244 uV, v_off_fin = 420.425 uV 11:25:01 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 92.274 uVs 11:25:01 Integrated Voltage Measurement... 11:25:10 Voltage Offset Correction: v_off_ini = 400.824 uV, v_off_fin = 413.584 uV 11:25:10 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -73.658 uVs 11:25:10 Integrated Voltage Measurement... 11:25:19 Voltage Offset Correction: v_off_ini = 415.498 uV, v_off_fin = 419.204 uV 11:25:19 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 108.220 uVs 11:25:19 Integrated Voltage Measurement... 11:25:28 Voltage Offset Correction: v_off_ini = 408.157 uV, v_off_fin = 410.075 uV 11:25:28 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -88.346 uVs 11:25:28 Integrated Voltage Measurement... 11:25:37 Voltage Offset Correction: v_off_ini = 419.849 uV, v_off_fin = 421.373 uV 11:25:37 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 76.658 uVs 11:25:41 I1X Average and RMS of Measurements VT = -87.354 +- 3.559 uVs I1X = 17.471 +- 0.712 uTm 11:25:44 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:25:48 Integrated Voltage Measurement... 11:25:57 Voltage Offset Correction: v_off_ini = 412.957 uV, v_off_fin = 407.907 uV 11:25:57 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -206.946 uVs 11:25:57 Integrated Voltage Measurement... 11:26:06 Voltage Offset Correction: v_off_ini = 421.271 uV, v_off_fin = 442.941 uV 11:26:06 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 221.874 uVs 11:26:06 Integrated Voltage Measurement... 11:26:15 Voltage Offset Correction: v_off_ini = 427.422 uV, v_off_fin = 425.331 uV 11:26:15 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -211.148 uVs 11:26:15 Integrated Voltage Measurement... 11:26:24 Voltage Offset Correction: v_off_ini = 429.656 uV, v_off_fin = 440.517 uV 11:26:24 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 235.138 uVs 11:26:24 Integrated Voltage Measurement... 11:26:33 Voltage Offset Correction: v_off_ini = 433.623 uV, v_off_fin = 414.635 uV 11:26:33 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -207.372 uVs 11:26:33 Integrated Voltage Measurement... 11:26:42 Voltage Offset Correction: v_off_ini = 421.896 uV, v_off_fin = 439.530 uV 11:26:42 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 220.184 uVs 11:26:46 I1Y Average and RMS of Measurements VT = -217.110 +- 4.273 uVs I1Y = -43.419 +- 0.853 uTm 11:26:51 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:26:54 Integrated Voltage Measurement... 11:27:03 Voltage Offset Correction: v_off_ini = 438.018 uV, v_off_fin = 430.903 uV 11:27:03 Integrated Voltage Measurement dXYus = -1.0002 mm, dXYds = 0.0000 mm dXY = -1.0002 mm VT = -67.968 uVs 11:27:03 Integrated Voltage Measurement... 11:27:12 Voltage Offset Correction: v_off_ini = 428.373 uV, v_off_fin = 452.350 uV 11:27:12 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 55.826 uVs 11:27:12 Integrated Voltage Measurement... 11:27:21 Voltage Offset Correction: v_off_ini = 424.280 uV, v_off_fin = 445.306 uV 11:27:21 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -58.632 uVs 11:27:21 Integrated Voltage Measurement... 11:27:30 Voltage Offset Correction: v_off_ini = 439.984 uV, v_off_fin = 462.700 uV 11:27:30 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 73.832 uVs 11:27:30 Integrated Voltage Measurement... 11:27:39 Voltage Offset Correction: v_off_ini = 446.476 uV, v_off_fin = 450.280 uV 11:27:39 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -79.590 uVs 11:27:39 Integrated Voltage Measurement... 11:27:48 Voltage Offset Correction: v_off_ini = 450.421 uV, v_off_fin = 455.840 uV 11:27:48 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 58.709 uVs 11:27:50 I2X Average and RMS of Measurements VT = -65.759 +- 2.979 uVs I2X = 6.050 +- 0.274 uTm^2 11:27:56 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:27:58 Integrated Voltage Measurement... 11:28:07 Voltage Offset Correction: v_off_ini = 451.093 uV, v_off_fin = 443.381 uV 11:28:07 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -406.711 uVs 11:28:07 Integrated Voltage Measurement... 11:28:16 Voltage Offset Correction: v_off_ini = 448.132 uV, v_off_fin = 441.801 uV 11:28:16 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 393.589 uVs 11:28:16 Integrated Voltage Measurement... 11:28:25 Voltage Offset Correction: v_off_ini = 442.986 uV, v_off_fin = 450.093 uV 11:28:25 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -382.361 uVs 11:28:25 Integrated Voltage Measurement... 11:28:34 Voltage Offset Correction: v_off_ini = 451.790 uV, v_off_fin = 450.889 uV 11:28:34 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 393.891 uVs 11:28:34 Integrated Voltage Measurement... 11:28:43 Voltage Offset Correction: v_off_ini = 444.898 uV, v_off_fin = 437.399 uV 11:28:43 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -391.044 uVs 11:28:43 Integrated Voltage Measurement... 11:28:52 Voltage Offset Correction: v_off_ini = 460.801 uV, v_off_fin = 462.806 uV 11:28:52 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 413.286 uVs 11:28:55 I2Y Average and RMS of Measurements VT = -396.813 +- 6.198 uVs I2Y = -36.507 +- 0.570 uTm^2 11:29:05 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:29:10 Integrated Voltage Measurement... 11:29:19 Voltage Offset Correction: v_off_ini = 455.342 uV, v_off_fin = 452.230 uV 11:29:19 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -104.759 uVs 11:29:19 Integrated Voltage Measurement... 11:29:28 Voltage Offset Correction: v_off_ini = 443.213 uV, v_off_fin = 449.856 uV 11:29:28 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 98.112 uVs 11:29:28 Integrated Voltage Measurement... 11:29:37 Voltage Offset Correction: v_off_ini = 464.278 uV, v_off_fin = 452.716 uV 11:29:37 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -86.037 uVs 11:29:37 Integrated Voltage Measurement... 11:29:46 Voltage Offset Correction: v_off_ini = 458.685 uV, v_off_fin = 453.859 uV 11:29:46 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 99.949 uVs 11:29:46 Integrated Voltage Measurement... 11:29:55 Voltage Offset Correction: v_off_ini = 456.605 uV, v_off_fin = 457.427 uV 11:29:55 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -75.333 uVs 11:29:55 Integrated Voltage Measurement... 11:30:04 Voltage Offset Correction: v_off_ini = 479.301 uV, v_off_fin = 474.013 uV 11:30:04 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 95.645 uVs 11:30:08 I1X Average and RMS of Measurements VT = -93.306 +- 6.514 uVs I1X = 18.662 +- 1.303 uTm 11:30:10 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:30:15 Integrated Voltage Measurement... 11:30:24 Voltage Offset Correction: v_off_ini = 470.914 uV, v_off_fin = 451.749 uV 11:30:24 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = -193.436 uVs 11:30:24 Integrated Voltage Measurement... 11:30:33 Voltage Offset Correction: v_off_ini = 462.547 uV, v_off_fin = 482.200 uV 11:30:33 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 209.813 uVs 11:30:33 Integrated Voltage Measurement... 11:30:42 Voltage Offset Correction: v_off_ini = 473.802 uV, v_off_fin = 461.662 uV 11:30:42 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -196.246 uVs 11:30:42 Integrated Voltage Measurement... 11:30:51 Voltage Offset Correction: v_off_ini = 462.053 uV, v_off_fin = 475.582 uV 11:30:51 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 201.037 uVs 11:30:51 Integrated Voltage Measurement... 11:31:00 Voltage Offset Correction: v_off_ini = 462.988 uV, v_off_fin = 460.683 uV 11:31:00 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -192.715 uVs 11:31:00 Integrated Voltage Measurement... 11:31:09 Voltage Offset Correction: v_off_ini = 471.379 uV, v_off_fin = 485.166 uV 11:31:09 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 196.179 uVs 11:31:13 I1Y Average and RMS of Measurements VT = -198.238 +- 2.944 uVs I1Y = -39.646 +- 0.590 uTm 11:31:18 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:31:21 Integrated Voltage Measurement... 11:31:30 Voltage Offset Correction: v_off_ini = 473.823 uV, v_off_fin = 473.677 uV 11:31:30 Integrated Voltage Measurement dXYus = -1.0002 mm, dXYds = 0.0000 mm dXY = -1.0002 mm VT = -98.709 uVs 11:31:30 Integrated Voltage Measurement... 11:31:39 Voltage Offset Correction: v_off_ini = 461.995 uV, v_off_fin = 478.492 uV 11:31:39 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 75.664 uVs 11:31:39 Integrated Voltage Measurement... 11:31:48 Voltage Offset Correction: v_off_ini = 481.688 uV, v_off_fin = 479.938 uV 11:31:48 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -68.685 uVs 11:31:48 Integrated Voltage Measurement... 11:31:57 Voltage Offset Correction: v_off_ini = 485.498 uV, v_off_fin = 487.767 uV 11:31:57 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 76.216 uVs 11:31:57 Integrated Voltage Measurement... 11:32:06 Voltage Offset Correction: v_off_ini = 476.964 uV, v_off_fin = 470.438 uV 11:32:06 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -82.272 uVs 11:32:06 Integrated Voltage Measurement... 11:32:15 Voltage Offset Correction: v_off_ini = 478.387 uV, v_off_fin = 490.811 uV 11:32:15 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 71.874 uVs 11:32:17 I2X Average and RMS of Measurements VT = -78.903 +- 6.153 uVs I2X = 7.259 +- 0.566 uTm^2 11:32:22 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:32:25 Integrated Voltage Measurement... 11:32:34 Voltage Offset Correction: v_off_ini = 485.480 uV, v_off_fin = 483.902 uV 11:32:34 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -365.712 uVs 11:32:34 Integrated Voltage Measurement... 11:32:43 Voltage Offset Correction: v_off_ini = 476.822 uV, v_off_fin = 475.195 uV 11:32:43 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 382.856 uVs 11:32:43 Integrated Voltage Measurement... 11:32:52 Voltage Offset Correction: v_off_ini = 476.840 uV, v_off_fin = 488.417 uV 11:32:52 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -383.115 uVs 11:32:52 Integrated Voltage Measurement... 11:33:01 Voltage Offset Correction: v_off_ini = 479.638 uV, v_off_fin = 481.190 uV 11:33:01 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 384.304 uVs 11:33:01 Integrated Voltage Measurement... 11:33:10 Voltage Offset Correction: v_off_ini = 490.949 uV, v_off_fin = 494.322 uV 11:33:10 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -377.798 uVs 11:33:10 Integrated Voltage Measurement... 11:33:19 Voltage Offset Correction: v_off_ini = 474.004 uV, v_off_fin = 493.105 uV 11:33:19 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 379.444 uVs 11:33:22 I2Y Average and RMS of Measurements VT = -378.872 +- 3.852 uVs I2Y = -34.856 +- 0.354 uTm^2 11:33:32 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:33:37 Integrated Voltage Measurement... 11:33:45 Voltage Offset Correction: v_off_ini = 477.460 uV, v_off_fin = 502.527 uV 11:33:46 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -98.015 uVs 11:33:46 Integrated Voltage Measurement... 11:33:55 Voltage Offset Correction: v_off_ini = 518.880 uV, v_off_fin = 516.092 uV 11:33:55 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 76.629 uVs 11:33:55 Integrated Voltage Measurement... 11:34:04 Voltage Offset Correction: v_off_ini = 492.353 uV, v_off_fin = 508.064 uV 11:34:04 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -112.276 uVs 11:34:04 Integrated Voltage Measurement... 11:34:13 Voltage Offset Correction: v_off_ini = 490.165 uV, v_off_fin = 510.650 uV 11:34:13 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 115.650 uVs 11:34:13 Integrated Voltage Measurement... 11:34:22 Voltage Offset Correction: v_off_ini = 512.964 uV, v_off_fin = 505.074 uV 11:34:22 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -99.590 uVs 11:34:22 Integrated Voltage Measurement... 11:34:31 Voltage Offset Correction: v_off_ini = 514.363 uV, v_off_fin = 516.961 uV 11:34:31 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 98.641 uVs 11:34:35 I1X Average and RMS of Measurements VT = -100.133 +- 10.900 uVs I1X = 20.027 +- 2.180 uTm 11:34:37 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:34:42 Integrated Voltage Measurement... 11:34:51 Voltage Offset Correction: v_off_ini = 497.947 uV, v_off_fin = 499.366 uV 11:34:51 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -185.789 uVs 11:34:51 Integrated Voltage Measurement... 11:35:00 Voltage Offset Correction: v_off_ini = 501.279 uV, v_off_fin = 508.255 uV 11:35:00 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 200.238 uVs 11:35:00 Integrated Voltage Measurement... 11:35:09 Voltage Offset Correction: v_off_ini = 502.479 uV, v_off_fin = 492.519 uV 11:35:09 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -189.053 uVs 11:35:09 Integrated Voltage Measurement... 11:35:18 Voltage Offset Correction: v_off_ini = 497.295 uV, v_off_fin = 501.357 uV 11:35:18 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 185.848 uVs 11:35:18 Integrated Voltage Measurement... 11:35:27 Voltage Offset Correction: v_off_ini = 493.580 uV, v_off_fin = 483.719 uV 11:35:27 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -186.561 uVs 11:35:27 Integrated Voltage Measurement... 11:35:36 Voltage Offset Correction: v_off_ini = 496.375 uV, v_off_fin = 501.840 uV 11:35:36 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 180.698 uVs 11:35:40 I1Y Average and RMS of Measurements VT = -188.031 +- 3.853 uVs I1Y = -37.603 +- 0.771 uTm 11:35:45 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:35:48 Integrated Voltage Measurement... 11:35:56 Voltage Offset Correction: v_off_ini = 507.868 uV, v_off_fin = 500.070 uV 11:35:57 Integrated Voltage Measurement dXYus = -1.0002 mm, dXYds = 0.0000 mm dXY = -1.0002 mm VT = -106.525 uVs 11:35:57 Integrated Voltage Measurement... 11:36:06 Voltage Offset Correction: v_off_ini = 512.792 uV, v_off_fin = 519.038 uV 11:36:06 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 88.703 uVs 11:36:06 Integrated Voltage Measurement... 11:36:15 Voltage Offset Correction: v_off_ini = 507.053 uV, v_off_fin = 495.144 uV 11:36:15 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -98.611 uVs 11:36:15 Integrated Voltage Measurement... 11:36:24 Voltage Offset Correction: v_off_ini = 492.889 uV, v_off_fin = 499.643 uV 11:36:24 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 72.412 uVs 11:36:24 Integrated Voltage Measurement... 11:36:33 Voltage Offset Correction: v_off_ini = 494.466 uV, v_off_fin = 489.135 uV 11:36:33 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -111.942 uVs 11:36:33 Integrated Voltage Measurement... 11:36:42 Voltage Offset Correction: v_off_ini = 489.007 uV, v_off_fin = 510.194 uV 11:36:42 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 79.354 uVs 11:36:44 I2X Average and RMS of Measurements VT = -92.925 +- 5.303 uVs I2X = 8.549 +- 0.488 uTm^2 11:36:49 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:36:52 Integrated Voltage Measurement... 11:37:01 Voltage Offset Correction: v_off_ini = 511.300 uV, v_off_fin = 483.361 uV 11:37:01 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -373.806 uVs 11:37:01 Integrated Voltage Measurement... 11:37:10 Voltage Offset Correction: v_off_ini = 494.078 uV, v_off_fin = 490.915 uV 11:37:10 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 341.731 uVs 11:37:10 Integrated Voltage Measurement... 11:37:19 Voltage Offset Correction: v_off_ini = 508.793 uV, v_off_fin = 484.339 uV 11:37:19 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -365.427 uVs 11:37:19 Integrated Voltage Measurement... 11:37:28 Voltage Offset Correction: v_off_ini = 498.806 uV, v_off_fin = 502.660 uV 11:37:28 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 358.991 uVs 11:37:28 Integrated Voltage Measurement... 11:37:37 Voltage Offset Correction: v_off_ini = 498.252 uV, v_off_fin = 499.444 uV 11:37:37 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -375.250 uVs 11:37:37 Integrated Voltage Measurement... 11:37:46 Voltage Offset Correction: v_off_ini = 504.944 uV, v_off_fin = 513.363 uV 11:37:46 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 371.977 uVs 11:37:48 I2Y Average and RMS of Measurements VT = -364.530 +- 6.674 uVs I2Y = -33.537 +- 0.614 uTm^2 11:37:59 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:38:03 Integrated Voltage Measurement... 11:38:12 Voltage Offset Correction: v_off_ini = 512.231 uV, v_off_fin = 507.535 uV 11:38:12 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.4998 mm dXY = -0.4999 mm VT = -108.006 uVs 11:38:12 Integrated Voltage Measurement... 11:38:21 Voltage Offset Correction: v_off_ini = 509.067 uV, v_off_fin = 504.585 uV 11:38:21 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 117.921 uVs 11:38:21 Integrated Voltage Measurement... 11:38:30 Voltage Offset Correction: v_off_ini = 513.567 uV, v_off_fin = 499.313 uV 11:38:30 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -100.727 uVs 11:38:30 Integrated Voltage Measurement... 11:38:39 Voltage Offset Correction: v_off_ini = 505.885 uV, v_off_fin = 511.189 uV 11:38:39 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 105.298 uVs 11:38:39 Integrated Voltage Measurement... 11:38:48 Voltage Offset Correction: v_off_ini = 508.316 uV, v_off_fin = 505.273 uV 11:38:48 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -108.021 uVs 11:38:49 Integrated Voltage Measurement... 11:38:57 Voltage Offset Correction: v_off_ini = 510.896 uV, v_off_fin = 507.427 uV 11:38:57 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 103.987 uVs 11:39:02 I1X Average and RMS of Measurements VT = -107.326 +- 4.169 uVs I1X = 21.467 +- 0.836 uTm 11:39:04 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:39:09 Integrated Voltage Measurement... 11:39:17 Voltage Offset Correction: v_off_ini = 507.724 uV, v_off_fin = 504.054 uV 11:39:18 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -178.925 uVs 11:39:18 Integrated Voltage Measurement... 11:39:27 Voltage Offset Correction: v_off_ini = 514.801 uV, v_off_fin = 520.050 uV 11:39:27 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 183.549 uVs 11:39:27 Integrated Voltage Measurement... 11:39:36 Voltage Offset Correction: v_off_ini = 516.889 uV, v_off_fin = 514.494 uV 11:39:36 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -185.389 uVs 11:39:36 Integrated Voltage Measurement... 11:39:45 Voltage Offset Correction: v_off_ini = 494.193 uV, v_off_fin = 526.613 uV 11:39:45 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 180.007 uVs 11:39:45 Integrated Voltage Measurement... 11:39:54 Voltage Offset Correction: v_off_ini = 516.900 uV, v_off_fin = 495.324 uV 11:39:54 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -174.417 uVs 11:39:54 Integrated Voltage Measurement... 11:40:03 Voltage Offset Correction: v_off_ini = 504.274 uV, v_off_fin = 522.143 uV 11:40:03 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 189.242 uVs 11:40:07 I1Y Average and RMS of Measurements VT = -181.922 +- 0.600 uVs I1Y = -36.381 +- 0.120 uTm 11:40:12 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:40:14 Integrated Voltage Measurement... 11:40:23 Voltage Offset Correction: v_off_ini = 495.977 uV, v_off_fin = 488.636 uV 11:40:23 Integrated Voltage Measurement dXYus = -1.0002 mm, dXYds = 0.0000 mm dXY = -1.0002 mm VT = -86.310 uVs 11:40:24 Integrated Voltage Measurement... 11:40:33 Voltage Offset Correction: v_off_ini = 499.794 uV, v_off_fin = 526.542 uV 11:40:33 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 74.332 uVs 11:40:33 Integrated Voltage Measurement... 11:40:42 Voltage Offset Correction: v_off_ini = 512.692 uV, v_off_fin = 516.128 uV 11:40:42 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -103.222 uVs 11:40:42 Integrated Voltage Measurement... 11:40:51 Voltage Offset Correction: v_off_ini = 503.327 uV, v_off_fin = 518.846 uV 11:40:51 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 86.778 uVs 11:40:51 Integrated Voltage Measurement... 11:41:00 Voltage Offset Correction: v_off_ini = 506.474 uV, v_off_fin = 496.135 uV 11:41:00 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -92.151 uVs 11:41:00 Integrated Voltage Measurement... 11:41:09 Voltage Offset Correction: v_off_ini = 514.941 uV, v_off_fin = 524.857 uV 11:41:09 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 67.315 uVs 11:41:11 I2X Average and RMS of Measurements VT = -85.018 +- 7.062 uVs I2X = 7.822 +- 0.650 uTm^2 11:41:16 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:41:19 Integrated Voltage Measurement... 11:41:28 Voltage Offset Correction: v_off_ini = 510.753 uV, v_off_fin = 504.881 uV 11:41:28 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = -337.802 uVs 11:41:28 Integrated Voltage Measurement... 11:41:37 Voltage Offset Correction: v_off_ini = 508.294 uV, v_off_fin = 512.017 uV 11:41:37 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 330.374 uVs 11:41:37 Integrated Voltage Measurement... 11:41:46 Voltage Offset Correction: v_off_ini = 519.928 uV, v_off_fin = 516.329 uV 11:41:46 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -364.677 uVs 11:41:46 Integrated Voltage Measurement... 11:41:55 Voltage Offset Correction: v_off_ini = 513.591 uV, v_off_fin = 508.237 uV 11:41:55 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 347.952 uVs 11:41:55 Integrated Voltage Measurement... 11:42:04 Voltage Offset Correction: v_off_ini = 507.158 uV, v_off_fin = 500.333 uV 11:42:04 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -327.788 uVs 11:42:04 Integrated Voltage Measurement... 11:42:13 Voltage Offset Correction: v_off_ini = 506.469 uV, v_off_fin = 507.084 uV 11:42:13 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 341.631 uVs 11:42:15 I2Y Average and RMS of Measurements VT = -341.704 +- 10.334 uVs I2Y = -31.437 +- 0.950 uTm^2 11:42:26 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:42:30 Integrated Voltage Measurement... 11:42:39 Voltage Offset Correction: v_off_ini = 514.493 uV, v_off_fin = 516.355 uV 11:42:39 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -134.659 uVs 11:42:39 Integrated Voltage Measurement... 11:42:48 Voltage Offset Correction: v_off_ini = 519.133 uV, v_off_fin = 525.287 uV 11:42:48 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 92.924 uVs 11:42:48 Integrated Voltage Measurement... 11:42:57 Voltage Offset Correction: v_off_ini = 523.279 uV, v_off_fin = 521.931 uV 11:42:57 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -119.229 uVs 11:42:57 Integrated Voltage Measurement... 11:43:06 Voltage Offset Correction: v_off_ini = 511.832 uV, v_off_fin = 524.234 uV 11:43:06 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 107.008 uVs 11:43:06 Integrated Voltage Measurement... 11:43:15 Voltage Offset Correction: v_off_ini = 519.837 uV, v_off_fin = 524.854 uV 11:43:15 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -115.150 uVs 11:43:15 Integrated Voltage Measurement... 11:43:24 Voltage Offset Correction: v_off_ini = 530.938 uV, v_off_fin = 534.115 uV 11:43:24 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 88.533 uVs 11:43:29 I1X Average and RMS of Measurements VT = -109.584 +- 5.482 uVs I1X = 21.918 +- 1.097 uTm 11:43:31 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:43:35 Integrated Voltage Measurement... 11:43:44 Voltage Offset Correction: v_off_ini = 535.717 uV, v_off_fin = 508.821 uV 11:43:44 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -166.953 uVs 11:43:44 Integrated Voltage Measurement... 11:43:53 Voltage Offset Correction: v_off_ini = 530.950 uV, v_off_fin = 536.603 uV 11:43:53 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5002 mm dXY = -0.5001 mm VT = 167.355 uVs 11:43:53 Integrated Voltage Measurement... 11:44:02 Voltage Offset Correction: v_off_ini = 535.806 uV, v_off_fin = 515.431 uV 11:44:02 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5002 mm dXY = 0.5001 mm VT = -170.128 uVs 11:44:02 Integrated Voltage Measurement... 11:44:11 Voltage Offset Correction: v_off_ini = 530.712 uV, v_off_fin = 561.649 uV 11:44:11 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 161.700 uVs 11:44:11 Integrated Voltage Measurement... 11:44:20 Voltage Offset Correction: v_off_ini = 538.820 uV, v_off_fin = 518.570 uV 11:44:20 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -172.836 uVs 11:44:20 Integrated Voltage Measurement... 11:44:29 Voltage Offset Correction: v_off_ini = 528.215 uV, v_off_fin = 533.375 uV 11:44:29 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5002 mm dXY = -0.5001 mm VT = 182.623 uVs 11:44:34 I1Y Average and RMS of Measurements VT = -170.266 +- 5.302 uVs I1Y = -34.049 +- 1.060 uTm 11:44:39 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:44:41 Integrated Voltage Measurement... 11:44:50 Voltage Offset Correction: v_off_ini = 528.490 uV, v_off_fin = 514.629 uV 11:44:50 Integrated Voltage Measurement dXYus = -1.0002 mm, dXYds = 0.0000 mm dXY = -1.0002 mm VT = -96.550 uVs 11:44:50 Integrated Voltage Measurement... 11:44:59 Voltage Offset Correction: v_off_ini = 525.925 uV, v_off_fin = 534.498 uV 11:44:59 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 97.772 uVs 11:44:59 Integrated Voltage Measurement... 11:45:08 Voltage Offset Correction: v_off_ini = 536.254 uV, v_off_fin = 531.253 uV 11:45:08 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -104.633 uVs 11:45:08 Integrated Voltage Measurement... 11:45:17 Voltage Offset Correction: v_off_ini = 532.939 uV, v_off_fin = 548.695 uV 11:45:17 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 87.702 uVs 11:45:17 Integrated Voltage Measurement... 11:45:26 Voltage Offset Correction: v_off_ini = 544.364 uV, v_off_fin = 541.208 uV 11:45:26 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -113.187 uVs 11:45:27 Integrated Voltage Measurement... 11:45:35 Voltage Offset Correction: v_off_ini = 542.223 uV, v_off_fin = 552.511 uV 11:45:36 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 98.207 uVs 11:45:38 I2X Average and RMS of Measurements VT = -99.675 +- 4.277 uVs I2X = 9.170 +- 0.394 uTm^2 11:45:43 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:45:46 Integrated Voltage Measurement... 11:45:55 Voltage Offset Correction: v_off_ini = 541.554 uV, v_off_fin = 540.388 uV 11:45:55 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -330.405 uVs 11:45:55 Integrated Voltage Measurement... 11:46:04 Voltage Offset Correction: v_off_ini = 545.658 uV, v_off_fin = 551.105 uV 11:46:04 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 317.385 uVs 11:46:04 Integrated Voltage Measurement... 11:46:13 Voltage Offset Correction: v_off_ini = 543.277 uV, v_off_fin = 539.445 uV 11:46:13 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -319.095 uVs 11:46:13 Integrated Voltage Measurement... 11:46:22 Voltage Offset Correction: v_off_ini = 556.552 uV, v_off_fin = 550.597 uV 11:46:22 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 312.073 uVs 11:46:22 Integrated Voltage Measurement... 11:46:31 Voltage Offset Correction: v_off_ini = 558.808 uV, v_off_fin = 551.770 uV 11:46:31 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -331.960 uVs 11:46:31 Integrated Voltage Measurement... 11:46:40 Voltage Offset Correction: v_off_ini = 559.686 uV, v_off_fin = 562.834 uV 11:46:40 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 320.812 uVs 11:46:42 I2Y Average and RMS of Measurements VT = -321.955 +- 4.618 uVs I2Y = -29.620 +- 0.425 uTm^2 11:46:53 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:46:57 Integrated Voltage Measurement... 11:47:06 Voltage Offset Correction: v_off_ini = 552.713 uV, v_off_fin = 555.024 uV 11:47:06 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -101.583 uVs 11:47:06 Integrated Voltage Measurement... 11:47:15 Voltage Offset Correction: v_off_ini = 567.210 uV, v_off_fin = 564.772 uV 11:47:15 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 91.718 uVs 11:47:15 Integrated Voltage Measurement... 11:47:24 Voltage Offset Correction: v_off_ini = 565.113 uV, v_off_fin = 551.514 uV 11:47:24 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -125.463 uVs 11:47:24 Integrated Voltage Measurement... 11:47:33 Voltage Offset Correction: v_off_ini = 574.224 uV, v_off_fin = 567.129 uV 11:47:33 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 88.331 uVs 11:47:33 Integrated Voltage Measurement... 11:47:42 Voltage Offset Correction: v_off_ini = 547.442 uV, v_off_fin = 545.167 uV 11:47:42 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -96.798 uVs 11:47:42 Integrated Voltage Measurement... 11:47:51 Voltage Offset Correction: v_off_ini = 553.710 uV, v_off_fin = 560.143 uV 11:47:51 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 114.173 uVs 11:47:55 I1X Average and RMS of Measurements VT = -103.011 +- 4.534 uVs I1X = 20.603 +- 0.907 uTm 11:47:58 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:48:02 Integrated Voltage Measurement... 11:48:11 Voltage Offset Correction: v_off_ini = 555.689 uV, v_off_fin = 548.539 uV 11:48:11 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -149.316 uVs 11:48:11 Integrated Voltage Measurement... 11:48:20 Voltage Offset Correction: v_off_ini = 554.512 uV, v_off_fin = 578.982 uV 11:48:20 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 166.826 uVs 11:48:20 Integrated Voltage Measurement... 11:48:29 Voltage Offset Correction: v_off_ini = 569.432 uV, v_off_fin = 562.312 uV 11:48:29 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -153.091 uVs 11:48:29 Integrated Voltage Measurement... 11:48:38 Voltage Offset Correction: v_off_ini = 569.368 uV, v_off_fin = 584.778 uV 11:48:38 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 168.831 uVs 11:48:38 Integrated Voltage Measurement... 11:48:47 Voltage Offset Correction: v_off_ini = 573.964 uV, v_off_fin = 566.946 uV 11:48:47 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -159.007 uVs 11:48:47 Integrated Voltage Measurement... 11:48:56 Voltage Offset Correction: v_off_ini = 578.080 uV, v_off_fin = 591.889 uV 11:48:56 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 169.294 uVs 11:49:00 I1Y Average and RMS of Measurements VT = -161.061 +- 2.483 uVs I1Y = -32.209 +- 0.497 uTm 11:49:06 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:49:08 Integrated Voltage Measurement... 11:49:17 Voltage Offset Correction: v_off_ini = 560.729 uV, v_off_fin = 547.232 uV 11:49:17 Integrated Voltage Measurement dXYus = -1.0002 mm, dXYds = 0.0000 mm dXY = -1.0002 mm VT = -88.679 uVs 11:49:17 Integrated Voltage Measurement... 11:49:26 Voltage Offset Correction: v_off_ini = 554.357 uV, v_off_fin = 576.079 uV 11:49:26 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 104.805 uVs 11:49:26 Integrated Voltage Measurement... 11:49:35 Voltage Offset Correction: v_off_ini = 554.501 uV, v_off_fin = 561.829 uV 11:49:35 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -85.110 uVs 11:49:35 Integrated Voltage Measurement... 11:49:44 Voltage Offset Correction: v_off_ini = 564.763 uV, v_off_fin = 575.766 uV 11:49:44 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 99.844 uVs 11:49:44 Integrated Voltage Measurement... 11:49:53 Voltage Offset Correction: v_off_ini = 549.469 uV, v_off_fin = 552.252 uV 11:49:53 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -110.587 uVs 11:49:53 Integrated Voltage Measurement... 11:50:02 Voltage Offset Correction: v_off_ini = 572.049 uV, v_off_fin = 567.186 uV 11:50:02 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 90.700 uVs 11:50:05 I2X Average and RMS of Measurements VT = -96.621 +- 3.335 uVs I2X = 8.889 +- 0.307 uTm^2 11:50:10 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:50:12 Integrated Voltage Measurement... 11:50:21 Voltage Offset Correction: v_off_ini = 569.588 uV, v_off_fin = 554.563 uV 11:50:21 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -310.666 uVs 11:50:22 Integrated Voltage Measurement... 11:50:31 Voltage Offset Correction: v_off_ini = 553.946 uV, v_off_fin = 566.275 uV 11:50:31 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 310.580 uVs 11:50:31 Integrated Voltage Measurement... 11:50:40 Voltage Offset Correction: v_off_ini = 558.582 uV, v_off_fin = 543.126 uV 11:50:40 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -317.683 uVs 11:50:40 Integrated Voltage Measurement... 11:50:49 Voltage Offset Correction: v_off_ini = 565.562 uV, v_off_fin = 564.224 uV 11:50:49 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 303.411 uVs 11:50:49 Integrated Voltage Measurement... 11:50:58 Voltage Offset Correction: v_off_ini = 564.140 uV, v_off_fin = 565.904 uV 11:50:58 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -325.949 uVs 11:50:58 Integrated Voltage Measurement... 11:51:07 Voltage Offset Correction: v_off_ini = 559.772 uV, v_off_fin = 559.725 uV 11:51:07 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 318.640 uVs 11:51:09 I2Y Average and RMS of Measurements VT = -314.488 +- 5.520 uVs I2Y = -28.933 +- 0.508 uTm^2 11:51:20 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:51:24 Integrated Voltage Measurement... 11:51:33 Voltage Offset Correction: v_off_ini = 561.003 uV, v_off_fin = 546.872 uV 11:51:33 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.4999 mm dXY = -0.4999 mm VT = -133.586 uVs 11:51:33 Integrated Voltage Measurement... 11:51:42 Voltage Offset Correction: v_off_ini = 542.278 uV, v_off_fin = 546.522 uV 11:51:42 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 103.909 uVs 11:51:42 Integrated Voltage Measurement... 11:51:51 Voltage Offset Correction: v_off_ini = 544.284 uV, v_off_fin = 542.642 uV 11:51:51 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -114.292 uVs 11:51:51 Integrated Voltage Measurement... 11:52:00 Voltage Offset Correction: v_off_ini = 551.363 uV, v_off_fin = 551.505 uV 11:52:00 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 117.197 uVs 11:52:00 Integrated Voltage Measurement... 11:52:09 Voltage Offset Correction: v_off_ini = 537.888 uV, v_off_fin = 544.270 uV 11:52:09 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -108.394 uVs 11:52:09 Integrated Voltage Measurement... 11:52:18 Voltage Offset Correction: v_off_ini = 543.755 uV, v_off_fin = 551.453 uV 11:52:18 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 133.090 uVs 11:52:22 I1X Average and RMS of Measurements VT = -118.411 +- 2.054 uVs I1X = 23.683 +- 0.411 uTm 11:52:25 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:52:29 Integrated Voltage Measurement... 11:52:38 Voltage Offset Correction: v_off_ini = 540.532 uV, v_off_fin = 542.469 uV 11:52:38 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -166.176 uVs 11:52:38 Integrated Voltage Measurement... 11:52:47 Voltage Offset Correction: v_off_ini = 554.502 uV, v_off_fin = 568.772 uV 11:52:47 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 161.531 uVs 11:52:47 Integrated Voltage Measurement... 11:52:56 Voltage Offset Correction: v_off_ini = 560.175 uV, v_off_fin = 552.658 uV 11:52:56 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -148.347 uVs 11:52:56 Integrated Voltage Measurement... 11:53:05 Voltage Offset Correction: v_off_ini = 552.745 uV, v_off_fin = 571.907 uV 11:53:05 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 162.771 uVs 11:53:05 Integrated Voltage Measurement... 11:53:14 Voltage Offset Correction: v_off_ini = 552.417 uV, v_off_fin = 538.453 uV 11:53:14 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -139.152 uVs 11:53:14 Integrated Voltage Measurement... 11:53:23 Voltage Offset Correction: v_off_ini = 547.742 uV, v_off_fin = 556.351 uV 11:53:23 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 156.576 uVs 11:53:27 I1Y Average and RMS of Measurements VT = -155.759 +- 6.530 uVs I1Y = -31.150 +- 1.305 uTm 11:53:33 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:53:35 Integrated Voltage Measurement... 11:53:44 Voltage Offset Correction: v_off_ini = 545.522 uV, v_off_fin = 551.593 uV 11:53:44 Integrated Voltage Measurement dXYus = -1.0002 mm, dXYds = 0.0000 mm dXY = -1.0002 mm VT = -122.570 uVs 11:53:44 Integrated Voltage Measurement... 11:53:53 Voltage Offset Correction: v_off_ini = 559.654 uV, v_off_fin = 563.408 uV 11:53:53 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 89.481 uVs 11:53:53 Integrated Voltage Measurement... 11:54:02 Voltage Offset Correction: v_off_ini = 547.328 uV, v_off_fin = 545.698 uV 11:54:02 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -92.293 uVs 11:54:02 Integrated Voltage Measurement... 11:54:11 Voltage Offset Correction: v_off_ini = 565.606 uV, v_off_fin = 577.679 uV 11:54:11 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 108.583 uVs 11:54:11 Integrated Voltage Measurement... 11:54:20 Voltage Offset Correction: v_off_ini = 556.251 uV, v_off_fin = 553.603 uV 11:54:20 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -103.644 uVs 11:54:20 Integrated Voltage Measurement... 11:54:29 Voltage Offset Correction: v_off_ini = 558.669 uV, v_off_fin = 565.763 uV 11:54:29 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 93.604 uVs 11:54:32 I2X Average and RMS of Measurements VT = -101.696 +- 3.150 uVs I2X = 9.356 +- 0.289 uTm^2 11:54:37 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:54:39 Integrated Voltage Measurement... 11:54:48 Voltage Offset Correction: v_off_ini = 558.648 uV, v_off_fin = 555.938 uV 11:54:48 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -315.569 uVs 11:54:49 Integrated Voltage Measurement... 11:54:57 Voltage Offset Correction: v_off_ini = 562.785 uV, v_off_fin = 566.324 uV 11:54:58 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 294.970 uVs 11:54:58 Integrated Voltage Measurement... 11:55:07 Voltage Offset Correction: v_off_ini = 564.208 uV, v_off_fin = 557.873 uV 11:55:07 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -306.006 uVs 11:55:07 Integrated Voltage Measurement... 11:55:16 Voltage Offset Correction: v_off_ini = 565.617 uV, v_off_fin = 563.416 uV 11:55:16 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 301.801 uVs 11:55:16 Integrated Voltage Measurement... 11:55:25 Voltage Offset Correction: v_off_ini = 573.551 uV, v_off_fin = 572.711 uV 11:55:25 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -284.421 uVs 11:55:25 Integrated Voltage Measurement... 11:55:34 Voltage Offset Correction: v_off_ini = 565.288 uV, v_off_fin = 563.042 uV 11:55:34 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 296.162 uVs 11:55:36 I2Y Average and RMS of Measurements VT = -299.821 +- 6.762 uVs I2Y = -27.584 +- 0.622 uTm^2 11:55:47 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:55:51 Integrated Voltage Measurement... 11:56:00 Voltage Offset Correction: v_off_ini = 562.815 uV, v_off_fin = 556.969 uV 11:56:00 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -118.765 uVs 11:56:00 Integrated Voltage Measurement... 11:56:09 Voltage Offset Correction: v_off_ini = 559.095 uV, v_off_fin = 555.508 uV 11:56:09 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 108.368 uVs 11:56:09 Integrated Voltage Measurement... 11:56:18 Voltage Offset Correction: v_off_ini = 556.107 uV, v_off_fin = 550.880 uV 11:56:18 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -126.362 uVs 11:56:18 Integrated Voltage Measurement... 11:56:27 Voltage Offset Correction: v_off_ini = 555.598 uV, v_off_fin = 572.487 uV 11:56:27 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 104.394 uVs 11:56:27 Integrated Voltage Measurement... 11:56:36 Voltage Offset Correction: v_off_ini = 576.391 uV, v_off_fin = 559.690 uV 11:56:36 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -137.471 uVs 11:56:36 Integrated Voltage Measurement... 11:56:45 Voltage Offset Correction: v_off_ini = 566.786 uV, v_off_fin = 569.160 uV 11:56:45 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 118.773 uVs 11:56:49 I1X Average and RMS of Measurements VT = -119.022 +- 6.477 uVs I1X = 23.805 +- 1.295 uTm 11:56:52 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:56:56 Integrated Voltage Measurement... 11:57:05 Voltage Offset Correction: v_off_ini = 569.797 uV, v_off_fin = 558.716 uV 11:57:05 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -159.065 uVs 11:57:05 Integrated Voltage Measurement... 11:57:14 Voltage Offset Correction: v_off_ini = 568.731 uV, v_off_fin = 571.532 uV 11:57:14 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 160.851 uVs 11:57:14 Integrated Voltage Measurement... 11:57:23 Voltage Offset Correction: v_off_ini = 554.459 uV, v_off_fin = 542.603 uV 11:57:23 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -150.521 uVs 11:57:23 Integrated Voltage Measurement... 11:57:32 Voltage Offset Correction: v_off_ini = 556.486 uV, v_off_fin = 572.717 uV 11:57:32 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 144.907 uVs 11:57:32 Integrated Voltage Measurement... 11:57:41 Voltage Offset Correction: v_off_ini = 555.202 uV, v_off_fin = 552.149 uV 11:57:41 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -153.883 uVs 11:57:41 Integrated Voltage Measurement... 11:57:50 Voltage Offset Correction: v_off_ini = 556.229 uV, v_off_fin = 567.512 uV 11:57:50 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5002 mm dXY = -0.5001 mm VT = 158.178 uVs 11:57:54 I1Y Average and RMS of Measurements VT = -154.567 +- 5.104 uVs I1Y = -30.911 +- 1.019 uTm 11:58:00 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:58:02 Integrated Voltage Measurement... 11:58:11 Voltage Offset Correction: v_off_ini = 568.911 uV, v_off_fin = 560.949 uV 11:58:11 Integrated Voltage Measurement dXYus = -1.0002 mm, dXYds = 0.0000 mm dXY = -1.0002 mm VT = -113.360 uVs 11:58:11 Integrated Voltage Measurement... 11:58:20 Voltage Offset Correction: v_off_ini = 564.135 uV, v_off_fin = 581.889 uV 11:58:20 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 105.852 uVs 11:58:20 Integrated Voltage Measurement... 11:58:29 Voltage Offset Correction: v_off_ini = 562.383 uV, v_off_fin = 545.415 uV 11:58:29 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -89.460 uVs 11:58:29 Integrated Voltage Measurement... 11:58:38 Voltage Offset Correction: v_off_ini = 554.758 uV, v_off_fin = 575.862 uV 11:58:38 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 99.038 uVs 11:58:38 Integrated Voltage Measurement... 11:58:47 Voltage Offset Correction: v_off_ini = 574.111 uV, v_off_fin = 562.668 uV 11:58:47 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -116.771 uVs 11:58:47 Integrated Voltage Measurement... 11:58:56 Voltage Offset Correction: v_off_ini = 573.014 uV, v_off_fin = 567.516 uV 11:58:56 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 109.945 uVs 11:58:59 I2X Average and RMS of Measurements VT = -105.738 +- 8.267 uVs I2X = 9.728 +- 0.760 uTm^2 11:59:04 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:59:06 Integrated Voltage Measurement... 11:59:15 Voltage Offset Correction: v_off_ini = 575.931 uV, v_off_fin = 568.362 uV 11:59:16 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -296.138 uVs 11:59:16 Integrated Voltage Measurement... 11:59:25 Voltage Offset Correction: v_off_ini = 563.917 uV, v_off_fin = 569.199 uV 11:59:25 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 283.296 uVs 11:59:25 Integrated Voltage Measurement... 11:59:34 Voltage Offset Correction: v_off_ini = 586.164 uV, v_off_fin = 576.312 uV 11:59:34 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -298.219 uVs 11:59:34 Integrated Voltage Measurement... 11:59:43 Voltage Offset Correction: v_off_ini = 561.318 uV, v_off_fin = 563.411 uV 11:59:43 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 310.146 uVs 11:59:43 Integrated Voltage Measurement... 11:59:52 Voltage Offset Correction: v_off_ini = 566.958 uV, v_off_fin = 561.327 uV 11:59:52 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -290.039 uVs 11:59:52 Integrated Voltage Measurement... 12:00:01 Voltage Offset Correction: v_off_ini = 562.026 uV, v_off_fin = 573.758 uV 12:00:01 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 275.094 uVs 12:00:03 I2Y Average and RMS of Measurements VT = -292.155 +- 8.991 uVs I2Y = -26.878 +- 0.827 uTm^2 12:00:16 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 12:00:20 Integrated Voltage Measurement... 12:00:29 Voltage Offset Correction: v_off_ini = 548.676 uV, v_off_fin = 544.284 uV 12:00:29 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -132.370 uVs 12:00:29 Integrated Voltage Measurement... 12:00:38 Voltage Offset Correction: v_off_ini = 566.367 uV, v_off_fin = 556.959 uV 12:00:38 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 126.515 uVs 12:00:38 Integrated Voltage Measurement... 12:00:47 Voltage Offset Correction: v_off_ini = 554.175 uV, v_off_fin = 559.877 uV 12:00:47 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -133.541 uVs 12:00:47 Integrated Voltage Measurement... 12:00:56 Voltage Offset Correction: v_off_ini = 563.255 uV, v_off_fin = 565.742 uV 12:00:56 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 131.437 uVs 12:00:56 Integrated Voltage Measurement... 12:01:05 Voltage Offset Correction: v_off_ini = 552.180 uV, v_off_fin = 550.486 uV 12:01:05 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -136.729 uVs 12:01:05 Integrated Voltage Measurement... 12:01:14 Voltage Offset Correction: v_off_ini = 560.987 uV, v_off_fin = 562.472 uV 12:01:14 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 130.015 uVs 12:01:19 I1X Average and RMS of Measurements VT = -131.768 +- 1.684 uVs I1X = 26.355 +- 0.337 uTm 12:01:21 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 12:01:25 Integrated Voltage Measurement... 12:01:34 Voltage Offset Correction: v_off_ini = 557.747 uV, v_off_fin = 554.992 uV 12:01:34 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -113.188 uVs 12:01:34 Integrated Voltage Measurement... 12:01:43 Voltage Offset Correction: v_off_ini = 564.407 uV, v_off_fin = 563.672 uV 12:01:43 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5002 mm dXY = -0.5001 mm VT = 151.739 uVs 12:01:43 Integrated Voltage Measurement... 12:01:52 Voltage Offset Correction: v_off_ini = 568.873 uV, v_off_fin = 553.558 uV 12:01:52 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5002 mm dXY = 0.5001 mm VT = -144.074 uVs 12:01:52 Integrated Voltage Measurement... 12:02:01 Voltage Offset Correction: v_off_ini = 564.701 uV, v_off_fin = 579.822 uV 12:02:01 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5002 mm dXY = -0.5001 mm VT = 160.866 uVs 12:02:01 Integrated Voltage Measurement... 12:02:10 Voltage Offset Correction: v_off_ini = 571.614 uV, v_off_fin = 557.796 uV 12:02:10 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -156.324 uVs 12:02:10 Integrated Voltage Measurement... 12:02:19 Voltage Offset Correction: v_off_ini = 579.812 uV, v_off_fin = 578.144 uV 12:02:19 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 133.907 uVs 12:02:24 I1Y Average and RMS of Measurements VT = -143.350 +- 8.262 uVs I1Y = -28.667 +- 1.652 uTm 12:02:29 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 12:02:31 Integrated Voltage Measurement... 12:02:40 Voltage Offset Correction: v_off_ini = 560.274 uV, v_off_fin = 564.280 uV 12:02:40 Integrated Voltage Measurement dXYus = -1.0002 mm, dXYds = 0.0000 mm dXY = -1.0002 mm VT = -138.517 uVs 12:02:40 Integrated Voltage Measurement... 12:02:49 Voltage Offset Correction: v_off_ini = 577.427 uV, v_off_fin = 584.629 uV 12:02:49 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 102.243 uVs 12:02:49 Integrated Voltage Measurement... 12:02:58 Voltage Offset Correction: v_off_ini = 587.863 uV, v_off_fin = 566.840 uV 12:02:58 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -112.973 uVs 12:02:58 Integrated Voltage Measurement... 12:03:07 Voltage Offset Correction: v_off_ini = 592.687 uV, v_off_fin = 590.350 uV 12:03:07 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 111.685 uVs 12:03:07 Integrated Voltage Measurement... 12:03:16 Voltage Offset Correction: v_off_ini = 587.051 uV, v_off_fin = 566.518 uV 12:03:16 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -125.953 uVs 12:03:16 Integrated Voltage Measurement... 12:03:25 Voltage Offset Correction: v_off_ini = 594.684 uV, v_off_fin = 587.156 uV 12:03:25 Integrated Voltage Measurement dXYus = 0.9998 mm, dXYds = 0.0000 mm dXY = 0.9998 mm VT = 112.084 uVs 12:03:28 I2X Average and RMS of Measurements VT = -117.243 +- 3.519 uVs I2X = 10.786 +- 0.324 uTm^2 12:03:33 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 12:03:36 Integrated Voltage Measurement... 12:03:45 Voltage Offset Correction: v_off_ini = 584.602 uV, v_off_fin = 584.382 uV 12:03:45 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -262.833 uVs 12:03:45 Integrated Voltage Measurement... 12:03:54 Voltage Offset Correction: v_off_ini = 587.724 uV, v_off_fin = 597.141 uV 12:03:54 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 258.955 uVs 12:03:54 Integrated Voltage Measurement... 12:04:03 Voltage Offset Correction: v_off_ini = 607.885 uV, v_off_fin = 614.115 uV 12:04:03 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -281.820 uVs 12:04:03 Integrated Voltage Measurement... 12:04:12 Voltage Offset Correction: v_off_ini = 595.392 uV, v_off_fin = 600.796 uV 12:04:12 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 273.759 uVs 12:04:12 Integrated Voltage Measurement... 12:04:21 Voltage Offset Correction: v_off_ini = 598.982 uV, v_off_fin = 587.052 uV 12:04:21 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -259.193 uVs 12:04:21 Integrated Voltage Measurement... 12:04:30 Voltage Offset Correction: v_off_ini = 603.400 uV, v_off_fin = 589.940 uV 12:04:30 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 268.028 uVs 12:04:32 I2Y Average and RMS of Measurements VT = -267.431 +- 7.408 uVs I2Y = -24.604 +- 0.682 uTm^2 12:04:44 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 12:04:48 Integrated Voltage Measurement... 12:04:57 Voltage Offset Correction: v_off_ini = 578.786 uV, v_off_fin = 581.059 uV 12:04:57 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -129.283 uVs 12:04:57 Integrated Voltage Measurement... 12:05:06 Voltage Offset Correction: v_off_ini = 580.100 uV, v_off_fin = 579.940 uV 12:05:06 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 149.356 uVs 12:05:06 Integrated Voltage Measurement... 12:05:15 Voltage Offset Correction: v_off_ini = 587.539 uV, v_off_fin = 580.024 uV 12:05:15 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -132.920 uVs 12:05:15 Integrated Voltage Measurement... 12:05:24 Voltage Offset Correction: v_off_ini = 597.946 uV, v_off_fin = 601.719 uV 12:05:24 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 147.176 uVs 12:05:24 Integrated Voltage Measurement... 12:05:33 Voltage Offset Correction: v_off_ini = 595.158 uV, v_off_fin = 598.364 uV 12:05:33 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -104.218 uVs 12:05:33 Integrated Voltage Measurement... 12:05:42 Voltage Offset Correction: v_off_ini = 603.358 uV, v_off_fin = 598.733 uV 12:05:42 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 132.050 uVs 12:05:47 I1X Average and RMS of Measurements VT = -132.500 +- 10.163 uVs I1X = 26.501 +- 2.033 uTm 12:05:49 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 12:05:53 Integrated Voltage Measurement... 12:06:02 Voltage Offset Correction: v_off_ini = 583.612 uV, v_off_fin = 588.433 uV 12:06:02 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -146.214 uVs 12:06:02 Integrated Voltage Measurement... 12:06:11 Voltage Offset Correction: v_off_ini = 591.286 uV, v_off_fin = 604.439 uV 12:06:11 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5002 mm dXY = -0.5001 mm VT = 155.185 uVs 12:06:12 Integrated Voltage Measurement... 12:06:20 Voltage Offset Correction: v_off_ini = 587.363 uV, v_off_fin = 569.249 uV 12:06:20 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5002 mm dXY = 0.5001 mm VT = -137.085 uVs 12:06:21 Integrated Voltage Measurement... 12:06:29 Voltage Offset Correction: v_off_ini = 576.967 uV, v_off_fin = 595.535 uV 12:06:29 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5002 mm dXY = -0.5001 mm VT = 129.880 uVs 12:06:30 Integrated Voltage Measurement... 12:06:38 Voltage Offset Correction: v_off_ini = 581.087 uV, v_off_fin = 581.091 uV 12:06:38 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -142.616 uVs 12:06:39 Integrated Voltage Measurement... 12:06:47 Voltage Offset Correction: v_off_ini = 600.518 uV, v_off_fin = 602.760 uV 12:06:47 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 132.462 uVs 12:06:52 I1Y Average and RMS of Measurements VT = -140.574 +- 7.349 uVs I1Y = -28.111 +- 1.470 uTm 12:06:57 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 12:06:59 Integrated Voltage Measurement... 12:07:08 Voltage Offset Correction: v_off_ini = 572.426 uV, v_off_fin = 579.814 uV 12:07:08 Integrated Voltage Measurement dXYus = -1.0002 mm, dXYds = 0.0000 mm dXY = -1.0002 mm VT = -128.042 uVs 12:07:08 Integrated Voltage Measurement... 12:07:17 Voltage Offset Correction: v_off_ini = 583.806 uV, v_off_fin = 585.319 uV 12:07:17 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 129.656 uVs 12:07:18 Integrated Voltage Measurement... 12:07:26 Voltage Offset Correction: v_off_ini = 582.099 uV, v_off_fin = 577.909 uV 12:07:27 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -124.535 uVs 12:07:27 Integrated Voltage Measurement... 12:07:36 Voltage Offset Correction: v_off_ini = 590.720 uV, v_off_fin = 592.033 uV 12:07:36 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 108.695 uVs 12:07:36 Integrated Voltage Measurement... 12:07:45 Voltage Offset Correction: v_off_ini = 579.381 uV, v_off_fin = 572.813 uV 12:07:45 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -127.221 uVs 12:07:45 Integrated Voltage Measurement... 12:07:54 Voltage Offset Correction: v_off_ini = 579.709 uV, v_off_fin = 589.934 uV 12:07:54 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 118.933 uVs 12:07:56 I2X Average and RMS of Measurements VT = -122.847 +- 4.997 uVs I2X = 11.302 +- 0.459 uTm^2 12:08:01 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 12:08:04 Integrated Voltage Measurement... 12:08:13 Voltage Offset Correction: v_off_ini = 580.893 uV, v_off_fin = 581.442 uV 12:08:13 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -249.174 uVs 12:08:13 Integrated Voltage Measurement... 12:08:22 Voltage Offset Correction: v_off_ini = 581.029 uV, v_off_fin = 587.718 uV 12:08:22 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 281.711 uVs 12:08:22 Integrated Voltage Measurement... 12:08:31 Voltage Offset Correction: v_off_ini = 591.689 uV, v_off_fin = 571.222 uV 12:08:31 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -263.503 uVs 12:08:31 Integrated Voltage Measurement... 12:08:40 Voltage Offset Correction: v_off_ini = 573.684 uV, v_off_fin = 571.165 uV 12:08:40 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 258.308 uVs 12:08:40 Integrated Voltage Measurement... 12:08:49 Voltage Offset Correction: v_off_ini = 567.292 uV, v_off_fin = 577.327 uV 12:08:49 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -264.895 uVs 12:08:49 Integrated Voltage Measurement... 12:08:58 Voltage Offset Correction: v_off_ini = 570.305 uV, v_off_fin = 566.104 uV 12:08:58 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 257.565 uVs 12:09:00 I2Y Average and RMS of Measurements VT = -262.526 +- 2.067 uVs I2Y = -24.152 +- 0.190 uTm^2 12:09:14 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 12:09:19 Integrated Voltage Measurement... 12:09:28 Voltage Offset Correction: v_off_ini = 585.933 uV, v_off_fin = 567.769 uV 12:09:28 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -137.996 uVs 12:09:28 Integrated Voltage Measurement... 12:09:37 Voltage Offset Correction: v_off_ini = 579.256 uV, v_off_fin = 581.208 uV 12:09:37 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 140.725 uVs 12:09:37 Integrated Voltage Measurement... 12:09:46 Voltage Offset Correction: v_off_ini = 583.640 uV, v_off_fin = 566.283 uV 12:09:46 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -145.869 uVs 12:09:46 Integrated Voltage Measurement... 12:09:55 Voltage Offset Correction: v_off_ini = 568.728 uV, v_off_fin = 578.809 uV 12:09:55 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 133.151 uVs 12:09:55 Integrated Voltage Measurement... 12:10:04 Voltage Offset Correction: v_off_ini = 564.604 uV, v_off_fin = 579.633 uV 12:10:04 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -120.953 uVs 12:10:04 Integrated Voltage Measurement... 12:10:13 Voltage Offset Correction: v_off_ini = 566.776 uV, v_off_fin = 577.415 uV 12:10:13 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 129.131 uVs 12:10:17 I1X Average and RMS of Measurements VT = -134.638 +- 6.785 uVs I1X = 26.928 +- 1.357 uTm 12:10:20 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 12:10:24 Integrated Voltage Measurement... 12:10:33 Voltage Offset Correction: v_off_ini = 585.505 uV, v_off_fin = 573.420 uV 12:10:33 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = -148.565 uVs 12:10:33 Integrated Voltage Measurement... 12:10:42 Voltage Offset Correction: v_off_ini = 567.539 uV, v_off_fin = 593.518 uV 12:10:42 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5002 mm dXY = -0.5001 mm VT = 146.947 uVs 12:10:42 Integrated Voltage Measurement... 12:10:51 Voltage Offset Correction: v_off_ini = 585.136 uV, v_off_fin = 567.485 uV 12:10:51 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5002 mm dXY = 0.5001 mm VT = -135.662 uVs 12:10:51 Integrated Voltage Measurement... 12:11:00 Voltage Offset Correction: v_off_ini = 571.089 uV, v_off_fin = 599.093 uV 12:11:00 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5002 mm dXY = -0.5001 mm VT = 146.607 uVs 12:11:00 Integrated Voltage Measurement... 12:11:09 Voltage Offset Correction: v_off_ini = 595.628 uV, v_off_fin = 563.839 uV 12:11:09 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5002 mm dXY = 0.5001 mm VT = -139.196 uVs 12:11:09 Integrated Voltage Measurement... 12:11:18 Voltage Offset Correction: v_off_ini = 578.053 uV, v_off_fin = 596.504 uV 12:11:18 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5002 mm dXY = -0.5001 mm VT = 148.192 uVs 12:11:22 I1Y Average and RMS of Measurements VT = -144.195 +- 2.726 uVs I1Y = -28.835 +- 0.547 uTm 12:11:27 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 12:11:30 Integrated Voltage Measurement... 12:11:39 Voltage Offset Correction: v_off_ini = 612.833 uV, v_off_fin = 591.365 uV 12:11:39 Integrated Voltage Measurement dXYus = -1.0002 mm, dXYds = 0.0000 mm dXY = -1.0002 mm VT = -144.320 uVs 12:11:39 Integrated Voltage Measurement... 12:11:48 Voltage Offset Correction: v_off_ini = 595.494 uV, v_off_fin = 599.332 uV 12:11:48 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 137.520 uVs 12:11:48 Integrated Voltage Measurement... 12:11:57 Voltage Offset Correction: v_off_ini = 594.233 uV, v_off_fin = 586.888 uV 12:11:57 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -131.432 uVs 12:11:57 Integrated Voltage Measurement... 12:12:06 Voltage Offset Correction: v_off_ini = 593.462 uV, v_off_fin = 603.076 uV 12:12:06 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 112.325 uVs 12:12:06 Integrated Voltage Measurement... 12:12:15 Voltage Offset Correction: v_off_ini = 602.300 uV, v_off_fin = 607.604 uV 12:12:15 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -136.800 uVs 12:12:15 Integrated Voltage Measurement... 12:12:24 Voltage Offset Correction: v_off_ini = 619.714 uV, v_off_fin = 622.993 uV 12:12:24 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 122.956 uVs 12:12:26 I2X Average and RMS of Measurements VT = -130.892 +- 7.807 uVs I2X = 12.042 +- 0.717 uTm^2 12:12:31 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 12:12:34 Integrated Voltage Measurement... 12:12:43 Voltage Offset Correction: v_off_ini = 605.843 uV, v_off_fin = 616.580 uV 12:12:43 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -256.720 uVs 12:12:43 Integrated Voltage Measurement... 12:12:52 Voltage Offset Correction: v_off_ini = 611.136 uV, v_off_fin = 605.832 uV 12:12:52 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 235.803 uVs 12:12:52 Integrated Voltage Measurement... 12:13:01 Voltage Offset Correction: v_off_ini = 608.478 uV, v_off_fin = 600.071 uV 12:13:01 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -243.681 uVs 12:13:01 Integrated Voltage Measurement... 12:13:10 Voltage Offset Correction: v_off_ini = 602.074 uV, v_off_fin = 598.915 uV 12:13:10 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 252.053 uVs 12:13:10 Integrated Voltage Measurement... 12:13:19 Voltage Offset Correction: v_off_ini = 608.998 uV, v_off_fin = 600.794 uV 12:13:19 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -241.493 uVs 12:13:19 Integrated Voltage Measurement... 12:13:28 Voltage Offset Correction: v_off_ini = 604.530 uV, v_off_fin = 603.708 uV 12:13:28 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 237.289 uVs 12:13:31 I2Y Average and RMS of Measurements VT = -244.506 +- 3.676 uVs I2Y = -22.495 +- 0.338 uTm^2 12:13:45 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 12:13:49 Integrated Voltage Measurement... 12:13:58 Voltage Offset Correction: v_off_ini = 630.216 uV, v_off_fin = 633.900 uV 12:13:58 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -135.184 uVs 12:13:58 Integrated Voltage Measurement... 12:14:07 Voltage Offset Correction: v_off_ini = 629.974 uV, v_off_fin = 618.713 uV 12:14:07 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 114.514 uVs 12:14:07 Integrated Voltage Measurement... 12:14:16 Voltage Offset Correction: v_off_ini = 611.392 uV, v_off_fin = 618.298 uV 12:14:16 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -111.850 uVs 12:14:16 Integrated Voltage Measurement... 12:14:25 Voltage Offset Correction: v_off_ini = 623.709 uV, v_off_fin = 627.935 uV 12:14:25 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 134.432 uVs 12:14:25 Integrated Voltage Measurement... 12:14:34 Voltage Offset Correction: v_off_ini = 624.383 uV, v_off_fin = 622.743 uV 12:14:34 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -129.660 uVs 12:14:34 Integrated Voltage Measurement... 12:14:43 Voltage Offset Correction: v_off_ini = 623.565 uV, v_off_fin = 616.947 uV 12:14:43 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 143.722 uVs 12:14:47 I1X Average and RMS of Measurements VT = -128.227 +- 6.025 uVs I1X = 25.646 +- 1.205 uTm 12:14:50 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 12:14:54 Integrated Voltage Measurement... 12:15:03 Voltage Offset Correction: v_off_ini = 635.521 uV, v_off_fin = 621.446 uV 12:15:03 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = -158.375 uVs 12:15:03 Integrated Voltage Measurement... 12:15:12 Voltage Offset Correction: v_off_ini = 636.076 uV, v_off_fin = 654.650 uV 12:15:12 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 154.638 uVs 12:15:12 Integrated Voltage Measurement... 12:15:21 Voltage Offset Correction: v_off_ini = 639.919 uV, v_off_fin = 619.171 uV 12:15:21 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5002 mm dXY = 0.5001 mm VT = -159.619 uVs 12:15:21 Integrated Voltage Measurement... 12:15:30 Voltage Offset Correction: v_off_ini = 650.294 uV, v_off_fin = 649.246 uV 12:15:30 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 144.800 uVs 12:15:30 Integrated Voltage Measurement... 12:15:39 Voltage Offset Correction: v_off_ini = 642.728 uV, v_off_fin = 624.495 uV 12:15:39 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -145.395 uVs 12:15:39 Integrated Voltage Measurement... 12:15:48 Voltage Offset Correction: v_off_ini = 637.759 uV, v_off_fin = 642.702 uV 12:15:48 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 166.701 uVs 12:15:52 I1Y Average and RMS of Measurements VT = -154.921 +- 1.927 uVs I1Y = -30.982 +- 0.387 uTm 12:15:57 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 12:16:00 Integrated Voltage Measurement... 12:16:09 Voltage Offset Correction: v_off_ini = 638.128 uV, v_off_fin = 637.262 uV 12:16:09 Integrated Voltage Measurement dXYus = -1.0002 mm, dXYds = 0.0000 mm dXY = -1.0002 mm VT = -125.568 uVs 12:16:09 Integrated Voltage Measurement... 12:16:18 Voltage Offset Correction: v_off_ini = 648.337 uV, v_off_fin = 665.799 uV 12:16:18 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 131.062 uVs 12:16:18 Integrated Voltage Measurement... 12:16:27 Voltage Offset Correction: v_off_ini = 641.669 uV, v_off_fin = 633.525 uV 12:16:27 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -122.508 uVs 12:16:27 Integrated Voltage Measurement... 12:16:36 Voltage Offset Correction: v_off_ini = 629.744 uV, v_off_fin = 653.241 uV 12:16:36 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 123.069 uVs 12:16:36 Integrated Voltage Measurement... 12:16:45 Voltage Offset Correction: v_off_ini = 631.742 uV, v_off_fin = 636.723 uV 12:16:45 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -76.726 uVs 12:16:45 Integrated Voltage Measurement... 12:16:54 Voltage Offset Correction: v_off_ini = 636.978 uV, v_off_fin = 651.131 uV 12:16:54 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 120.755 uVs 12:16:57 I2X Average and RMS of Measurements VT = -116.615 +- 12.839 uVs I2X = 10.729 +- 1.181 uTm^2 12:17:02 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 12:17:04 Integrated Voltage Measurement... 12:17:13 Voltage Offset Correction: v_off_ini = 654.684 uV, v_off_fin = 648.230 uV 12:17:13 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -251.227 uVs 12:17:13 Integrated Voltage Measurement... 12:17:22 Voltage Offset Correction: v_off_ini = 639.412 uV, v_off_fin = 642.970 uV 12:17:23 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 259.463 uVs 12:17:23 Integrated Voltage Measurement... 12:17:32 Voltage Offset Correction: v_off_ini = 635.128 uV, v_off_fin = 642.259 uV 12:17:32 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -248.237 uVs 12:17:32 Integrated Voltage Measurement... 12:17:41 Voltage Offset Correction: v_off_ini = 649.077 uV, v_off_fin = 634.340 uV 12:17:41 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 234.637 uVs 12:17:41 Integrated Voltage Measurement... 12:17:50 Voltage Offset Correction: v_off_ini = 632.406 uV, v_off_fin = 626.392 uV 12:17:50 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -229.166 uVs 12:17:50 Integrated Voltage Measurement... 12:17:59 Voltage Offset Correction: v_off_ini = 637.337 uV, v_off_fin = 643.772 uV 12:17:59 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 233.757 uVs 12:18:01 I2Y Average and RMS of Measurements VT = -242.748 +- 9.794 uVs I2Y = -22.333 +- 0.901 uTm^2 12:18:15 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 12:18:19 Integrated Voltage Measurement... 12:18:28 Voltage Offset Correction: v_off_ini = 643.898 uV, v_off_fin = 625.357 uV 12:18:28 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -143.453 uVs 12:18:28 Integrated Voltage Measurement... 12:18:37 Voltage Offset Correction: v_off_ini = 624.102 uV, v_off_fin = 613.132 uV 12:18:37 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 106.014 uVs 12:18:37 Integrated Voltage Measurement... 12:18:46 Voltage Offset Correction: v_off_ini = 606.517 uV, v_off_fin = 610.649 uV 12:18:46 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -105.768 uVs 12:18:46 Integrated Voltage Measurement... 12:18:55 Voltage Offset Correction: v_off_ini = 617.108 uV, v_off_fin = 624.690 uV 12:18:55 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 116.544 uVs 12:18:56 Integrated Voltage Measurement... 12:19:04 Voltage Offset Correction: v_off_ini = 615.839 uV, v_off_fin = 613.697 uV 12:19:05 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -103.626 uVs 12:19:05 Integrated Voltage Measurement... 12:19:14 Voltage Offset Correction: v_off_ini = 621.456 uV, v_off_fin = 617.826 uV 12:19:14 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 119.270 uVs 12:19:18 I1X Average and RMS of Measurements VT = -115.779 +- 6.333 uVs I1X = 23.156 +- 1.267 uTm 12:19:20 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 12:19:25 Integrated Voltage Measurement... 12:19:34 Voltage Offset Correction: v_off_ini = 614.112 uV, v_off_fin = 596.752 uV 12:19:34 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = -153.710 uVs 12:19:34 Integrated Voltage Measurement... 12:19:42 Voltage Offset Correction: v_off_ini = 605.061 uV, v_off_fin = 625.156 uV 12:19:43 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 174.207 uVs 12:19:43 Integrated Voltage Measurement... 12:19:52 Voltage Offset Correction: v_off_ini = 613.530 uV, v_off_fin = 615.339 uV 12:19:52 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -155.091 uVs 12:19:52 Integrated Voltage Measurement... 12:20:01 Voltage Offset Correction: v_off_ini = 617.020 uV, v_off_fin = 629.651 uV 12:20:01 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 162.234 uVs 12:20:01 Integrated Voltage Measurement... 12:20:10 Voltage Offset Correction: v_off_ini = 621.227 uV, v_off_fin = 612.935 uV 12:20:10 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -174.386 uVs 12:20:10 Integrated Voltage Measurement... 12:20:19 Voltage Offset Correction: v_off_ini = 619.355 uV, v_off_fin = 640.488 uV 12:20:19 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5002 mm dXY = -0.5001 mm VT = 167.676 uVs 12:20:23 I1Y Average and RMS of Measurements VT = -164.551 +- 5.067 uVs I1Y = -32.907 +- 1.012 uTm 12:20:28 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 12:20:30 Integrated Voltage Measurement... 12:20:39 Voltage Offset Correction: v_off_ini = 623.612 uV, v_off_fin = 622.244 uV 12:20:40 Integrated Voltage Measurement dXYus = -1.0001 mm, dXYds = 0.0000 mm dXY = -1.0001 mm VT = -110.558 uVs 12:20:40 Integrated Voltage Measurement... 12:20:49 Voltage Offset Correction: v_off_ini = 624.229 uV, v_off_fin = 639.833 uV 12:20:49 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 99.520 uVs 12:20:49 Integrated Voltage Measurement... 12:20:58 Voltage Offset Correction: v_off_ini = 617.561 uV, v_off_fin = 622.408 uV 12:20:58 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -111.557 uVs 12:20:58 Integrated Voltage Measurement... 12:21:07 Voltage Offset Correction: v_off_ini = 643.937 uV, v_off_fin = 652.922 uV 12:21:07 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 104.019 uVs 12:21:07 Integrated Voltage Measurement... 12:21:16 Voltage Offset Correction: v_off_ini = 631.016 uV, v_off_fin = 615.965 uV 12:21:16 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -101.794 uVs 12:21:16 Integrated Voltage Measurement... 12:21:25 Voltage Offset Correction: v_off_ini = 632.141 uV, v_off_fin = 644.604 uV 12:21:25 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 111.702 uVs 12:21:27 I2X Average and RMS of Measurements VT = -106.525 +- 1.133 uVs I2X = 9.800 +- 0.104 uTm^2 12:21:32 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 12:21:35 Integrated Voltage Measurement... 12:21:44 Voltage Offset Correction: v_off_ini = 657.562 uV, v_off_fin = 643.020 uV 12:21:44 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -262.769 uVs 12:21:44 Integrated Voltage Measurement... 12:21:53 Voltage Offset Correction: v_off_ini = 632.297 uV, v_off_fin = 632.416 uV 12:21:53 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 252.335 uVs 12:21:53 Integrated Voltage Measurement... 12:22:02 Voltage Offset Correction: v_off_ini = 635.373 uV, v_off_fin = 636.367 uV 12:22:02 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -241.479 uVs 12:22:02 Integrated Voltage Measurement... 12:22:11 Voltage Offset Correction: v_off_ini = 614.969 uV, v_off_fin = 627.852 uV 12:22:11 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 248.256 uVs 12:22:11 Integrated Voltage Measurement... 12:22:20 Voltage Offset Correction: v_off_ini = 627.538 uV, v_off_fin = 628.108 uV 12:22:20 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -251.797 uVs 12:22:20 Integrated Voltage Measurement... 12:22:29 Voltage Offset Correction: v_off_ini = 630.480 uV, v_off_fin = 636.428 uV 12:22:29 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 237.795 uVs 12:22:31 I2Y Average and RMS of Measurements VT = -249.072 +- 5.996 uVs I2Y = -22.915 +- 0.552 uTm^2 12:22:46 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 12:22:50 Integrated Voltage Measurement... 12:22:59 Voltage Offset Correction: v_off_ini = 636.158 uV, v_off_fin = 623.168 uV 12:22:59 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -113.742 uVs 12:22:59 Integrated Voltage Measurement... 12:23:08 Voltage Offset Correction: v_off_ini = 748.066 uV, v_off_fin = 620.864 uV 12:23:08 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 183.791 uVs 12:23:08 Integrated Voltage Measurement... 12:23:17 Voltage Offset Correction: v_off_ini = 733.010 uV, v_off_fin = 637.272 uV 12:23:17 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -224.041 uVs 12:23:17 Integrated Voltage Measurement... 12:23:26 Voltage Offset Correction: v_off_ini = 619.137 uV, v_off_fin = 634.886 uV 12:23:26 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 430.618 uVs 12:23:26 Integrated Voltage Measurement... 12:23:35 Voltage Offset Correction: v_off_ini = 835.838 uV, v_off_fin = 633.405 uV 12:23:35 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -340.615 uVs 12:23:35 Integrated Voltage Measurement... 12:23:44 Voltage Offset Correction: v_off_ini = 634.362 uV, v_off_fin = 635.168 uV 12:23:44 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 132.991 uVs 12:23:48 I1X Average and RMS of Measurements VT = -237.633 +- 72.900 uVs I1X = 47.527 +- 14.580 uTm 12:23:51 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 12:23:55 Integrated Voltage Measurement... 12:24:04 Voltage Offset Correction: v_off_ini = 626.678 uV, v_off_fin = 611.919 uV 12:24:04 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = -160.388 uVs 12:24:04 Integrated Voltage Measurement... 12:24:13 Voltage Offset Correction: v_off_ini = 639.529 uV, v_off_fin = 636.724 uV 12:24:13 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 197.385 uVs 12:24:13 Integrated Voltage Measurement... 12:24:22 Voltage Offset Correction: v_off_ini = 637.577 uV, v_off_fin = 627.089 uV 12:24:22 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -182.311 uVs 12:24:22 Integrated Voltage Measurement... 12:24:31 Voltage Offset Correction: v_off_ini = 638.227 uV, v_off_fin = 649.028 uV 12:24:31 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 167.697 uVs 12:24:31 Integrated Voltage Measurement... 12:24:40 Voltage Offset Correction: v_off_ini = 638.579 uV, v_off_fin = 627.530 uV 12:24:40 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -182.263 uVs 12:24:40 Integrated Voltage Measurement... 12:24:49 Voltage Offset Correction: v_off_ini = 635.806 uV, v_off_fin = 651.989 uV 12:24:49 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 187.890 uVs 12:24:53 I1Y Average and RMS of Measurements VT = -179.656 +- 4.148 uVs I1Y = -35.929 +- 0.830 uTm 12:24:58 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 12:25:01 Integrated Voltage Measurement... 12:25:10 Voltage Offset Correction: v_off_ini = 647.494 uV, v_off_fin = 638.296 uV 12:25:10 Integrated Voltage Measurement dXYus = -1.0002 mm, dXYds = 0.0000 mm dXY = -1.0002 mm VT = -115.440 uVs 12:25:10 Integrated Voltage Measurement... 12:25:19 Voltage Offset Correction: v_off_ini = 635.511 uV, v_off_fin = 657.283 uV 12:25:19 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 105.421 uVs 12:25:19 Integrated Voltage Measurement... 12:25:28 Voltage Offset Correction: v_off_ini = 649.000 uV, v_off_fin = 636.498 uV 12:25:28 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -120.499 uVs 12:25:28 Integrated Voltage Measurement... 12:25:37 Voltage Offset Correction: v_off_ini = 637.332 uV, v_off_fin = 644.498 uV 12:25:37 Integrated Voltage Measurement dXYus = 0.9998 mm, dXYds = 0.0000 mm dXY = 0.9998 mm VT = 132.641 uVs 12:25:37 Integrated Voltage Measurement... 12:25:46 Voltage Offset Correction: v_off_ini = 638.922 uV, v_off_fin = 642.076 uV 12:25:46 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -120.595 uVs 12:25:46 Integrated Voltage Measurement... 12:25:55 Voltage Offset Correction: v_off_ini = 651.274 uV, v_off_fin = 654.394 uV 12:25:55 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 84.815 uVs 12:25:58 I2X Average and RMS of Measurements VT = -113.235 +- 9.943 uVs I2X = 10.418 +- 0.915 uTm^2 12:26:03 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 12:26:05 Integrated Voltage Measurement... 12:26:14 Voltage Offset Correction: v_off_ini = 643.644 uV, v_off_fin = 650.426 uV 12:26:14 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -249.967 uVs 12:26:14 Integrated Voltage Measurement... 12:26:23 Voltage Offset Correction: v_off_ini = 649.604 uV, v_off_fin = 652.392 uV 12:26:23 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 255.038 uVs 12:26:23 Integrated Voltage Measurement... 12:26:32 Voltage Offset Correction: v_off_ini = 646.214 uV, v_off_fin = 650.488 uV 12:26:32 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -243.807 uVs 12:26:32 Integrated Voltage Measurement... 12:26:41 Voltage Offset Correction: v_off_ini = 653.069 uV, v_off_fin = 640.802 uV 12:26:41 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 229.910 uVs 12:26:42 Integrated Voltage Measurement... 12:26:50 Voltage Offset Correction: v_off_ini = 636.077 uV, v_off_fin = 646.578 uV 12:26:51 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -247.045 uVs 12:26:51 Integrated Voltage Measurement... 12:27:00 Voltage Offset Correction: v_off_ini = 626.226 uV, v_off_fin = 629.994 uV 12:27:00 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 267.416 uVs 12:27:02 I2Y Average and RMS of Measurements VT = -248.864 +- 8.706 uVs I2Y = -22.895 +- 0.801 uTm^2 12:27:16 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 12:27:20 Integrated Voltage Measurement... 12:27:29 Voltage Offset Correction: v_off_ini = 655.529 uV, v_off_fin = 646.951 uV 12:27:29 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -129.389 uVs 12:27:29 Integrated Voltage Measurement... 12:27:38 Voltage Offset Correction: v_off_ini = 649.122 uV, v_off_fin = 632.289 uV 12:27:38 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 98.372 uVs 12:27:38 Integrated Voltage Measurement... 12:27:47 Voltage Offset Correction: v_off_ini = 658.436 uV, v_off_fin = 656.848 uV 12:27:47 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -106.632 uVs 12:27:47 Integrated Voltage Measurement... 12:27:56 Voltage Offset Correction: v_off_ini = 651.121 uV, v_off_fin = 658.662 uV 12:27:56 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 118.777 uVs 12:27:56 Integrated Voltage Measurement... 12:28:05 Voltage Offset Correction: v_off_ini = 655.952 uV, v_off_fin = 652.363 uV 12:28:05 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -96.361 uVs 12:28:05 Integrated Voltage Measurement... 12:28:14 Voltage Offset Correction: v_off_ini = 669.363 uV, v_off_fin = 672.326 uV 12:28:14 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 119.821 uVs 12:28:19 I1X Average and RMS of Measurements VT = -111.559 +- 2.499 uVs I1X = 22.312 +- 0.500 uTm 12:28:21 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 12:28:25 Integrated Voltage Measurement... 12:28:34 Voltage Offset Correction: v_off_ini = 661.038 uV, v_off_fin = 652.098 uV 12:28:34 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -190.001 uVs 12:28:34 Integrated Voltage Measurement... 12:28:43 Voltage Offset Correction: v_off_ini = 674.232 uV, v_off_fin = 672.187 uV 12:28:43 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 192.515 uVs 12:28:43 Integrated Voltage Measurement... 12:28:52 Voltage Offset Correction: v_off_ini = 655.093 uV, v_off_fin = 643.462 uV 12:28:52 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -179.447 uVs 12:28:52 Integrated Voltage Measurement... 12:29:01 Voltage Offset Correction: v_off_ini = 641.848 uV, v_off_fin = 650.665 uV 12:29:01 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 191.663 uVs 12:29:01 Integrated Voltage Measurement... 12:29:10 Voltage Offset Correction: v_off_ini = 649.667 uV, v_off_fin = 637.443 uV 12:29:10 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5002 mm dXY = 0.5001 mm VT = -192.157 uVs 12:29:10 Integrated Voltage Measurement... 12:29:19 Voltage Offset Correction: v_off_ini = 633.341 uV, v_off_fin = 662.904 uV 12:29:19 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 188.437 uVs 12:29:24 I1Y Average and RMS of Measurements VT = -189.037 +- 2.493 uVs I1Y = -37.804 +- 0.499 uTm 12:29:29 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 12:29:31 Integrated Voltage Measurement... 12:29:40 Voltage Offset Correction: v_off_ini = 671.173 uV, v_off_fin = 653.510 uV 12:29:40 Integrated Voltage Measurement dXYus = -1.0001 mm, dXYds = 0.0000 mm dXY = -1.0001 mm VT = -117.521 uVs 12:29:40 Integrated Voltage Measurement... 12:29:49 Voltage Offset Correction: v_off_ini = 664.426 uV, v_off_fin = 680.849 uV 12:29:49 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 98.558 uVs 12:29:49 Integrated Voltage Measurement... 12:29:58 Voltage Offset Correction: v_off_ini = 664.993 uV, v_off_fin = 649.680 uV 12:29:59 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -106.282 uVs 12:29:59 Integrated Voltage Measurement... 12:30:08 Voltage Offset Correction: v_off_ini = 655.487 uV, v_off_fin = 678.302 uV 12:30:08 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 90.435 uVs 12:30:08 Integrated Voltage Measurement... 12:30:17 Voltage Offset Correction: v_off_ini = 667.770 uV, v_off_fin = 661.392 uV 12:30:17 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -92.155 uVs 12:30:17 Integrated Voltage Measurement... 12:30:26 Voltage Offset Correction: v_off_ini = 671.269 uV, v_off_fin = 683.817 uV 12:30:26 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 101.240 uVs 12:30:28 I2X Average and RMS of Measurements VT = -101.032 +- 5.001 uVs I2X = 9.295 +- 0.460 uTm^2 12:30:33 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 12:30:36 Integrated Voltage Measurement... 12:30:45 Voltage Offset Correction: v_off_ini = 680.612 uV, v_off_fin = 672.994 uV 12:30:45 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = -256.924 uVs 12:30:45 Integrated Voltage Measurement... 12:30:54 Voltage Offset Correction: v_off_ini = 666.488 uV, v_off_fin = 670.180 uV 12:30:54 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 248.406 uVs 12:30:54 Integrated Voltage Measurement... 12:31:03 Voltage Offset Correction: v_off_ini = 676.468 uV, v_off_fin = 663.351 uV 12:31:03 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -255.306 uVs 12:31:03 Integrated Voltage Measurement... 12:31:12 Voltage Offset Correction: v_off_ini = 671.221 uV, v_off_fin = 670.425 uV 12:31:12 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 251.686 uVs 12:31:12 Integrated Voltage Measurement... 12:31:21 Voltage Offset Correction: v_off_ini = 672.020 uV, v_off_fin = 671.680 uV 12:31:21 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -251.922 uVs 12:31:21 Integrated Voltage Measurement... 12:31:30 Voltage Offset Correction: v_off_ini = 680.863 uV, v_off_fin = 678.163 uV 12:31:30 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 253.256 uVs 12:31:32 I2Y Average and RMS of Measurements VT = -252.917 +- 0.411 uVs I2Y = -23.269 +- 0.038 uTm^2 12:31:47 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 12:31:51 Integrated Voltage Measurement... 12:32:00 Voltage Offset Correction: v_off_ini = 681.339 uV, v_off_fin = 669.265 uV 12:32:00 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -106.216 uVs 12:32:00 Integrated Voltage Measurement... 12:32:09 Voltage Offset Correction: v_off_ini = 698.716 uV, v_off_fin = 672.532 uV 12:32:09 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 77.157 uVs 12:32:09 Integrated Voltage Measurement... 12:32:18 Voltage Offset Correction: v_off_ini = 672.684 uV, v_off_fin = 666.854 uV 12:32:18 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -113.458 uVs 12:32:18 Integrated Voltage Measurement... 12:32:27 Voltage Offset Correction: v_off_ini = 674.744 uV, v_off_fin = 676.372 uV 12:32:27 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 122.838 uVs 12:32:27 Integrated Voltage Measurement... 12:32:36 Voltage Offset Correction: v_off_ini = 679.799 uV, v_off_fin = 664.275 uV 12:32:36 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -112.747 uVs 12:32:36 Integrated Voltage Measurement... 12:32:45 Voltage Offset Correction: v_off_ini = 675.997 uV, v_off_fin = 684.180 uV 12:32:45 Integrated Voltage Measurement dXYus = 0.4998 mm, dXYds = 0.5000 mm dXY = 0.4999 mm VT = 67.458 uVs 12:32:49 I1X Average and RMS of Measurements VT = -99.979 +- 12.864 uVs I1X = 19.996 +- 2.572 uTm 12:32:52 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 12:32:56 Integrated Voltage Measurement... 12:33:05 Voltage Offset Correction: v_off_ini = 659.718 uV, v_off_fin = 640.965 uV 12:33:05 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -176.898 uVs 12:33:05 Integrated Voltage Measurement... 12:33:14 Voltage Offset Correction: v_off_ini = 668.593 uV, v_off_fin = 669.374 uV 12:33:14 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 216.821 uVs 12:33:14 Integrated Voltage Measurement... 12:33:23 Voltage Offset Correction: v_off_ini = 662.440 uV, v_off_fin = 641.533 uV 12:33:23 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -182.302 uVs 12:33:23 Integrated Voltage Measurement... 12:33:32 Voltage Offset Correction: v_off_ini = 662.437 uV, v_off_fin = 677.286 uV 12:33:32 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 201.561 uVs 12:33:32 Integrated Voltage Measurement... 12:33:41 Voltage Offset Correction: v_off_ini = 663.964 uV, v_off_fin = 658.918 uV 12:33:41 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -193.416 uVs 12:33:41 Integrated Voltage Measurement... 12:33:50 Voltage Offset Correction: v_off_ini = 672.683 uV, v_off_fin = 683.125 uV 12:33:50 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = 199.775 uVs 12:33:54 I1Y Average and RMS of Measurements VT = -195.129 +- 2.263 uVs I1Y = -39.024 +- 0.454 uTm 12:33:59 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 12:34:02 Integrated Voltage Measurement... 12:34:11 Voltage Offset Correction: v_off_ini = 685.210 uV, v_off_fin = 674.098 uV 12:34:11 Integrated Voltage Measurement dXYus = -1.0001 mm, dXYds = 0.0000 mm dXY = -1.0001 mm VT = -111.089 uVs 12:34:11 Integrated Voltage Measurement... 12:34:20 Voltage Offset Correction: v_off_ini = 681.061 uV, v_off_fin = 698.498 uV 12:34:20 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 100.469 uVs 12:34:20 Integrated Voltage Measurement... 12:34:29 Voltage Offset Correction: v_off_ini = 687.435 uV, v_off_fin = 678.551 uV 12:34:29 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -98.136 uVs 12:34:29 Integrated Voltage Measurement... 12:34:38 Voltage Offset Correction: v_off_ini = 683.674 uV, v_off_fin = 685.407 uV 12:34:38 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 107.876 uVs 12:34:38 Integrated Voltage Measurement... 12:34:47 Voltage Offset Correction: v_off_ini = 678.401 uV, v_off_fin = 673.781 uV 12:34:47 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -88.370 uVs 12:34:47 Integrated Voltage Measurement... 12:34:56 Voltage Offset Correction: v_off_ini = 701.909 uV, v_off_fin = 682.314 uV 12:34:56 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 106.692 uVs 12:34:59 I2X Average and RMS of Measurements VT = -102.105 +- 3.427 uVs I2X = 9.394 +- 0.315 uTm^2 12:35:04 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 12:35:06 Integrated Voltage Measurement... 12:35:15 Voltage Offset Correction: v_off_ini = 687.507 uV, v_off_fin = 680.993 uV 12:35:15 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -253.396 uVs 12:35:15 Integrated Voltage Measurement... 12:35:24 Voltage Offset Correction: v_off_ini = 681.816 uV, v_off_fin = 682.121 uV 12:35:24 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 259.440 uVs 12:35:24 Integrated Voltage Measurement... 12:35:33 Voltage Offset Correction: v_off_ini = 681.884 uV, v_off_fin = 679.951 uV 12:35:33 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -247.313 uVs 12:35:33 Integrated Voltage Measurement... 12:35:42 Voltage Offset Correction: v_off_ini = 672.163 uV, v_off_fin = 676.391 uV 12:35:43 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 236.760 uVs 12:35:43 Integrated Voltage Measurement... 12:35:52 Voltage Offset Correction: v_off_ini = 680.572 uV, v_off_fin = 679.006 uV 12:35:52 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -249.407 uVs 12:35:52 Integrated Voltage Measurement... 12:36:01 Voltage Offset Correction: v_off_ini = 683.615 uV, v_off_fin = 690.445 uV 12:36:01 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 245.548 uVs 12:36:03 I2Y Average and RMS of Measurements VT = -248.644 +- 5.929 uVs I2Y = -22.875 +- 0.545 uTm^2 12:36:06 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm