SLAC Magnetic Measurements Date: 07-24-2025 Time: 11:44:27 Project: LCLS-II-HE Magnet Type: Phase Shifter Magnet Name: HE_SXPS_16341 Dataset Number: 1 Run Purpose: Tuning Run Type: Stretched Wire Run Number: 3 Measurement Device: Operator: YL Comment: 11:44:35 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:44:37 Motion Parameters, X Controller, Axis 1 acc = 0.001000 m/s/s dcl = 0.001000 m/s/s vel = 0.001000 m/s pos = 0.000000 m kp = 188.600000 ki = 820.000000 kd = 491.400000 db = 0.000000 11:44:39 Motion Parameters, X Controller, Axis 2 acc = 0.001000 m/s/s dcl = 0.001000 m/s/s vel = 0.001000 m/s pos = 0.000000 m kp = 188.600000 ki = 820.000000 kd = 491.400000 db = 0.000000 11:44:40 Motion Parameters, Y Controller, Axis 1 acc = 0.001000 m/s/s dcl = 0.001000 m/s/s vel = 0.001000 m/s pos = 0.000000 m kp = 188.600000 ki = 820.000000 kd = 491.400000 db = 0.000000 11:44:42 Motion Parameters, Y Controller, Axis 2 acc = 0.001000 m/s/s dcl = 0.001000 m/s/s vel = 0.001000 m/s pos = 0.000000 m kp = 188.600000 ki = 820.000000 kd = 491.400000 db = 0.000000 11:45:19 Wire Position: X0 = -2.0000 mm, Y0 = 0.0000 mm 11:45:23 Integrated Voltage Measurement... 11:45:32 Voltage Offset Correction: v_off_ini = 381.745 uV, v_off_fin = 377.319 uV 11:45:32 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4998 mm dXY = -0.4999 mm VT = -56.110 uVs 11:45:32 Integrated Voltage Measurement... 11:45:41 Voltage Offset Correction: v_off_ini = 378.419 uV, v_off_fin = 381.688 uV 11:45:41 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 40.214 uVs 11:45:41 Integrated Voltage Measurement... 11:45:50 Voltage Offset Correction: v_off_ini = 382.474 uV, v_off_fin = 374.762 uV 11:45:50 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -15.191 uVs 11:45:50 Integrated Voltage Measurement... 11:45:59 Voltage Offset Correction: v_off_ini = 378.882 uV, v_off_fin = 375.465 uV 11:45:59 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 28.214 uVs 11:46:00 Integrated Voltage Measurement... 11:46:08 Voltage Offset Correction: v_off_ini = 375.139 uV, v_off_fin = 371.368 uV 11:46:08 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -44.591 uVs 11:46:09 Integrated Voltage Measurement... 11:46:17 Voltage Offset Correction: v_off_ini = 378.935 uV, v_off_fin = 387.608 uV 11:46:18 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.4999 mm dXY = 0.4999 mm VT = 48.410 uVs 11:46:22 I1X Average and RMS of Measurements VT = -38.788 +- 12.101 uVs I1X = 7.759 +- 2.421 uTm 11:46:24 Wire Position: X0 = -2.0000 mm, Y0 = 0.0000 mm 11:46:29 Integrated Voltage Measurement... 11:46:37 Voltage Offset Correction: v_off_ini = 361.873 uV, v_off_fin = 355.629 uV 11:46:38 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -272.079 uVs 11:46:38 Integrated Voltage Measurement... 11:46:46 Voltage Offset Correction: v_off_ini = 366.197 uV, v_off_fin = 387.090 uV 11:46:47 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 275.120 uVs 11:46:47 Integrated Voltage Measurement... 11:46:55 Voltage Offset Correction: v_off_ini = 375.006 uV, v_off_fin = 361.420 uV 11:46:56 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -290.906 uVs 11:46:56 Integrated Voltage Measurement... 11:47:04 Voltage Offset Correction: v_off_ini = 386.819 uV, v_off_fin = 375.819 uV 11:47:05 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 258.326 uVs 11:47:05 Integrated Voltage Measurement... 11:47:14 Voltage Offset Correction: v_off_ini = 366.412 uV, v_off_fin = 363.215 uV 11:47:14 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -268.930 uVs 11:47:14 Integrated Voltage Measurement... 11:47:23 Voltage Offset Correction: v_off_ini = 378.749 uV, v_off_fin = 386.454 uV 11:47:23 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 277.638 uVs 11:47:27 I1Y Average and RMS of Measurements VT = -273.833 +- 0.568 uVs I1Y = -54.767 +- 0.114 uTm 11:47:32 Wire Position: X0 = -2.0000 mm, Y0 = 0.0000 mm 11:47:34 Integrated Voltage Measurement... 11:47:43 Voltage Offset Correction: v_off_ini = 369.614 uV, v_off_fin = 374.721 uV 11:47:44 Integrated Voltage Measurement dXYus = -1.0002 mm, dXYds = 0.0000 mm dXY = -1.0002 mm VT = -30.147 uVs 11:47:44 Integrated Voltage Measurement... 11:47:53 Voltage Offset Correction: v_off_ini = 363.710 uV, v_off_fin = 395.540 uV 11:47:53 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 44.123 uVs 11:47:53 Integrated Voltage Measurement... 11:48:02 Voltage Offset Correction: v_off_ini = 378.328 uV, v_off_fin = 368.764 uV 11:48:02 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -45.656 uVs 11:48:02 Integrated Voltage Measurement... 11:48:11 Voltage Offset Correction: v_off_ini = 368.798 uV, v_off_fin = 376.940 uV 11:48:11 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 47.869 uVs 11:48:11 Integrated Voltage Measurement... 11:48:20 Voltage Offset Correction: v_off_ini = 373.006 uV, v_off_fin = 367.916 uV 11:48:20 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -45.125 uVs 11:48:20 Integrated Voltage Measurement... 11:48:29 Voltage Offset Correction: v_off_ini = 372.283 uV, v_off_fin = 383.470 uV 11:48:29 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 56.020 uVs 11:48:31 I2X Average and RMS of Measurements VT = -44.823 +- 5.655 uVs I2X = 4.124 +- 0.520 uTm^2 11:48:36 Wire Position: X0 = -2.0000 mm, Y0 = 0.0000 mm 11:48:39 Integrated Voltage Measurement... 11:48:48 Voltage Offset Correction: v_off_ini = 362.308 uV, v_off_fin = 364.930 uV 11:48:48 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -537.624 uVs 11:48:48 Integrated Voltage Measurement... 11:48:57 Voltage Offset Correction: v_off_ini = 356.278 uV, v_off_fin = 369.567 uV 11:48:57 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 556.381 uVs 11:48:57 Integrated Voltage Measurement... 11:49:06 Voltage Offset Correction: v_off_ini = 355.648 uV, v_off_fin = 369.006 uV 11:49:06 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -553.513 uVs 11:49:06 Integrated Voltage Measurement... 11:49:15 Voltage Offset Correction: v_off_ini = 359.661 uV, v_off_fin = 365.214 uV 11:49:15 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 556.097 uVs 11:49:15 Integrated Voltage Measurement... 11:49:24 Voltage Offset Correction: v_off_ini = 373.003 uV, v_off_fin = 358.832 uV 11:49:24 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -555.601 uVs 11:49:24 Integrated Voltage Measurement... 11:49:33 Voltage Offset Correction: v_off_ini = 356.184 uV, v_off_fin = 352.674 uV 11:49:33 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 545.622 uVs 11:49:36 I2Y Average and RMS of Measurements VT = -550.806 +- 3.188 uVs I2Y = -50.674 +- 0.293 uTm^2 11:49:40 Wire Position: X0 = -1.0000 mm, Y0 = 0.0000 mm 11:49:44 Integrated Voltage Measurement... 11:49:53 Voltage Offset Correction: v_off_ini = 359.565 uV, v_off_fin = 373.355 uV 11:49:53 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -73.228 uVs 11:49:53 Integrated Voltage Measurement... 11:50:02 Voltage Offset Correction: v_off_ini = 371.777 uV, v_off_fin = 355.355 uV 11:50:02 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 40.084 uVs 11:50:03 Integrated Voltage Measurement... 11:50:11 Voltage Offset Correction: v_off_ini = 365.075 uV, v_off_fin = 361.569 uV 11:50:12 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -48.936 uVs 11:50:12 Integrated Voltage Measurement... 11:50:20 Voltage Offset Correction: v_off_ini = 367.037 uV, v_off_fin = 366.179 uV 11:50:20 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 47.511 uVs 11:50:21 Integrated Voltage Measurement... 11:50:29 Voltage Offset Correction: v_off_ini = 377.099 uV, v_off_fin = 358.680 uV 11:50:29 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -43.766 uVs 11:50:29 Integrated Voltage Measurement... 11:50:38 Voltage Offset Correction: v_off_ini = 361.275 uV, v_off_fin = 374.515 uV 11:50:38 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 43.802 uVs 11:50:43 I1X Average and RMS of Measurements VT = -49.554 +- 5.339 uVs I1X = 9.911 +- 1.068 uTm 11:50:45 Wire Position: X0 = -1.0000 mm, Y0 = 0.0000 mm 11:50:49 Integrated Voltage Measurement... 11:50:58 Voltage Offset Correction: v_off_ini = 363.076 uV, v_off_fin = 371.992 uV 11:50:58 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -288.104 uVs 11:50:58 Integrated Voltage Measurement... 11:51:07 Voltage Offset Correction: v_off_ini = 361.348 uV, v_off_fin = 372.863 uV 11:51:07 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 290.312 uVs 11:51:07 Integrated Voltage Measurement... 11:51:16 Voltage Offset Correction: v_off_ini = 351.194 uV, v_off_fin = 343.973 uV 11:51:16 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -291.328 uVs 11:51:16 Integrated Voltage Measurement... 11:51:25 Voltage Offset Correction: v_off_ini = 369.157 uV, v_off_fin = 377.017 uV 11:51:25 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 272.450 uVs 11:51:25 Integrated Voltage Measurement... 11:51:34 Voltage Offset Correction: v_off_ini = 366.213 uV, v_off_fin = 367.477 uV 11:51:34 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -282.375 uVs 11:51:34 Integrated Voltage Measurement... 11:51:43 Voltage Offset Correction: v_off_ini = 370.632 uV, v_off_fin = 397.823 uV 11:51:43 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 298.221 uVs 11:51:47 I1Y Average and RMS of Measurements VT = -287.132 +- 3.734 uVs I1Y = -57.426 +- 0.747 uTm 11:51:53 Wire Position: X0 = -1.0000 mm, Y0 = 0.0000 mm 11:51:56 Integrated Voltage Measurement... 11:52:05 Voltage Offset Correction: v_off_ini = 366.265 uV, v_off_fin = 360.977 uV 11:52:05 Integrated Voltage Measurement dXYus = -1.0002 mm, dXYds = 0.0000 mm dXY = -1.0002 mm VT = -70.516 uVs 11:52:05 Integrated Voltage Measurement... 11:52:14 Voltage Offset Correction: v_off_ini = 381.555 uV, v_off_fin = 384.512 uV 11:52:14 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 53.618 uVs 11:52:14 Integrated Voltage Measurement... 11:52:23 Voltage Offset Correction: v_off_ini = 378.434 uV, v_off_fin = 364.780 uV 11:52:23 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -67.892 uVs 11:52:23 Integrated Voltage Measurement... 11:52:32 Voltage Offset Correction: v_off_ini = 387.412 uV, v_off_fin = 390.551 uV 11:52:32 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 70.817 uVs 11:52:32 Integrated Voltage Measurement... 11:52:41 Voltage Offset Correction: v_off_ini = 363.012 uV, v_off_fin = 360.680 uV 11:52:41 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -75.204 uVs 11:52:41 Integrated Voltage Measurement... 11:52:50 Voltage Offset Correction: v_off_ini = 369.300 uV, v_off_fin = 378.654 uV 11:52:50 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 58.117 uVs 11:52:52 I2X Average and RMS of Measurements VT = -66.028 +- 3.009 uVs I2X = 6.074 +- 0.277 uTm^2 11:52:57 Wire Position: X0 = -1.0000 mm, Y0 = 0.0000 mm 11:53:00 Integrated Voltage Measurement... 11:53:09 Voltage Offset Correction: v_off_ini = 375.229 uV, v_off_fin = 370.861 uV 11:53:09 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -542.365 uVs 11:53:09 Integrated Voltage Measurement... 11:53:18 Voltage Offset Correction: v_off_ini = 386.355 uV, v_off_fin = 393.300 uV 11:53:18 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 553.277 uVs 11:53:18 Integrated Voltage Measurement... 11:53:27 Voltage Offset Correction: v_off_ini = 392.519 uV, v_off_fin = 373.722 uV 11:53:27 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -563.074 uVs 11:53:27 Integrated Voltage Measurement... 11:53:36 Voltage Offset Correction: v_off_ini = 397.801 uV, v_off_fin = 397.370 uV 11:53:36 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 558.596 uVs 11:53:36 Integrated Voltage Measurement... 11:53:45 Voltage Offset Correction: v_off_ini = 395.310 uV, v_off_fin = 401.996 uV 11:53:45 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -565.109 uVs 11:53:45 Integrated Voltage Measurement... 11:53:54 Voltage Offset Correction: v_off_ini = 394.358 uV, v_off_fin = 401.962 uV 11:53:54 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 537.482 uVs 11:53:57 I2Y Average and RMS of Measurements VT = -553.317 +- 5.502 uVs I2Y = -50.905 +- 0.506 uTm^2 11:54:01 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:54:05 Integrated Voltage Measurement... 11:54:14 Voltage Offset Correction: v_off_ini = 403.944 uV, v_off_fin = 387.023 uV 11:54:14 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -57.720 uVs 11:54:14 Integrated Voltage Measurement... 11:54:23 Voltage Offset Correction: v_off_ini = 392.463 uV, v_off_fin = 391.325 uV 11:54:23 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 74.088 uVs 11:54:23 Integrated Voltage Measurement... 11:54:32 Voltage Offset Correction: v_off_ini = 399.221 uV, v_off_fin = 402.620 uV 11:54:33 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -89.449 uVs 11:54:33 Integrated Voltage Measurement... 11:54:42 Voltage Offset Correction: v_off_ini = 414.365 uV, v_off_fin = 418.901 uV 11:54:42 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 57.070 uVs 11:54:42 Integrated Voltage Measurement... 11:54:51 Voltage Offset Correction: v_off_ini = 404.792 uV, v_off_fin = 406.122 uV 11:54:51 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -68.582 uVs 11:54:51 Integrated Voltage Measurement... 11:55:00 Voltage Offset Correction: v_off_ini = 404.833 uV, v_off_fin = 411.192 uV 11:55:00 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 52.199 uVs 11:55:04 I1X Average and RMS of Measurements VT = -66.518 +- 5.272 uVs I1X = 13.304 +- 1.054 uTm 11:55:06 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:55:11 Integrated Voltage Measurement... 11:55:20 Voltage Offset Correction: v_off_ini = 401.690 uV, v_off_fin = 393.260 uV 11:55:20 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -286.405 uVs 11:55:20 Integrated Voltage Measurement... 11:55:29 Voltage Offset Correction: v_off_ini = 401.188 uV, v_off_fin = 415.531 uV 11:55:29 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 280.296 uVs 11:55:29 Integrated Voltage Measurement... 11:55:38 Voltage Offset Correction: v_off_ini = 404.114 uV, v_off_fin = 385.387 uV 11:55:38 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -284.823 uVs 11:55:38 Integrated Voltage Measurement... 11:55:47 Voltage Offset Correction: v_off_ini = 386.373 uV, v_off_fin = 411.764 uV 11:55:47 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 295.757 uVs 11:55:47 Integrated Voltage Measurement... 11:55:56 Voltage Offset Correction: v_off_ini = 397.071 uV, v_off_fin = 384.055 uV 11:55:56 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -273.172 uVs 11:55:56 Integrated Voltage Measurement... 11:56:05 Voltage Offset Correction: v_off_ini = 402.687 uV, v_off_fin = 420.379 uV 11:56:05 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 290.861 uVs 11:56:09 I1Y Average and RMS of Measurements VT = -285.219 +- 3.627 uVs I1Y = -57.044 +- 0.725 uTm 11:56:14 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:56:17 Integrated Voltage Measurement... 11:56:26 Voltage Offset Correction: v_off_ini = 399.189 uV, v_off_fin = 407.107 uV 11:56:26 Integrated Voltage Measurement dXYus = -1.0002 mm, dXYds = 0.0000 mm dXY = -1.0002 mm VT = -88.314 uVs 11:56:26 Integrated Voltage Measurement... 11:56:35 Voltage Offset Correction: v_off_ini = 421.435 uV, v_off_fin = 428.079 uV 11:56:35 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 67.583 uVs 11:56:35 Integrated Voltage Measurement... 11:56:44 Voltage Offset Correction: v_off_ini = 422.889 uV, v_off_fin = 421.133 uV 11:56:44 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -82.790 uVs 11:56:44 Integrated Voltage Measurement... 11:56:53 Voltage Offset Correction: v_off_ini = 426.333 uV, v_off_fin = 421.076 uV 11:56:53 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 73.035 uVs 11:56:53 Integrated Voltage Measurement... 11:57:02 Voltage Offset Correction: v_off_ini = 418.288 uV, v_off_fin = 403.454 uV 11:57:02 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -81.829 uVs 11:57:02 Integrated Voltage Measurement... 11:57:13 Voltage Offset Correction: v_off_ini = 412.415 uV, v_off_fin = 440.067 uV 11:57:13 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 69.465 uVs 11:57:16 I2X Average and RMS of Measurements VT = -77.169 +- 1.076 uVs I2X = 7.100 +- 0.099 uTm^2 11:57:21 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:57:24 Integrated Voltage Measurement... 11:57:33 Voltage Offset Correction: v_off_ini = 429.459 uV, v_off_fin = 417.792 uV 11:57:33 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -573.726 uVs 11:57:33 Integrated Voltage Measurement... 11:57:42 Voltage Offset Correction: v_off_ini = 414.588 uV, v_off_fin = 417.849 uV 11:57:42 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 549.618 uVs 11:57:42 Integrated Voltage Measurement... 11:57:51 Voltage Offset Correction: v_off_ini = 415.641 uV, v_off_fin = 415.832 uV 11:57:51 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -560.769 uVs 11:57:51 Integrated Voltage Measurement... 11:58:00 Voltage Offset Correction: v_off_ini = 418.507 uV, v_off_fin = 421.850 uV 11:58:00 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 549.632 uVs 11:58:00 Integrated Voltage Measurement... 11:58:09 Voltage Offset Correction: v_off_ini = 412.255 uV, v_off_fin = 409.271 uV 11:58:09 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -541.651 uVs 11:58:09 Integrated Voltage Measurement... 11:58:18 Voltage Offset Correction: v_off_ini = 431.289 uV, v_off_fin = 420.276 uV 11:58:18 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 552.787 uVs 11:58:20 I2Y Average and RMS of Measurements VT = -554.697 +- 5.911 uVs I2Y = -51.032 +- 0.544 uTm^2 11:58:25 Wire Position: X0 = 1.0000 mm, Y0 = 0.0000 mm 11:58:29 Integrated Voltage Measurement... 11:58:38 Voltage Offset Correction: v_off_ini = 424.088 uV, v_off_fin = 399.572 uV 11:58:38 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -99.890 uVs 11:58:38 Integrated Voltage Measurement... 11:58:47 Voltage Offset Correction: v_off_ini = 406.236 uV, v_off_fin = 417.767 uV 11:58:47 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 84.055 uVs 11:58:47 Integrated Voltage Measurement... 11:58:56 Voltage Offset Correction: v_off_ini = 423.902 uV, v_off_fin = 411.909 uV 11:58:56 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -84.344 uVs 11:58:57 Integrated Voltage Measurement... 11:59:05 Voltage Offset Correction: v_off_ini = 430.950 uV, v_off_fin = 429.588 uV 11:59:05 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 88.032 uVs 11:59:06 Integrated Voltage Measurement... 11:59:14 Voltage Offset Correction: v_off_ini = 431.517 uV, v_off_fin = 422.548 uV 11:59:14 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -93.185 uVs 11:59:15 Integrated Voltage Measurement... 11:59:23 Voltage Offset Correction: v_off_ini = 418.856 uV, v_off_fin = 415.146 uV 11:59:24 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 91.167 uVs 11:59:28 I1X Average and RMS of Measurements VT = -90.112 +- 2.776 uVs I1X = 18.023 +- 0.555 uTm 11:59:30 Wire Position: X0 = 1.0000 mm, Y0 = 0.0000 mm 11:59:35 Integrated Voltage Measurement... 11:59:43 Voltage Offset Correction: v_off_ini = 432.411 uV, v_off_fin = 423.693 uV 11:59:44 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -287.847 uVs 11:59:44 Integrated Voltage Measurement... 11:59:52 Voltage Offset Correction: v_off_ini = 442.566 uV, v_off_fin = 443.335 uV 11:59:53 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 296.506 uVs 11:59:53 Integrated Voltage Measurement... 12:00:02 Voltage Offset Correction: v_off_ini = 441.507 uV, v_off_fin = 427.149 uV 12:00:02 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -281.271 uVs 12:00:02 Integrated Voltage Measurement... 12:00:11 Voltage Offset Correction: v_off_ini = 438.195 uV, v_off_fin = 439.785 uV 12:00:11 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 295.645 uVs 12:00:11 Integrated Voltage Measurement... 12:00:20 Voltage Offset Correction: v_off_ini = 436.156 uV, v_off_fin = 430.771 uV 12:00:20 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -296.007 uVs 12:00:20 Integrated Voltage Measurement... 12:00:29 Voltage Offset Correction: v_off_ini = 433.148 uV, v_off_fin = 451.399 uV 12:00:29 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 295.355 uVs 12:00:33 I1Y Average and RMS of Measurements VT = -292.105 +- 2.949 uVs I1Y = -58.421 +- 0.590 uTm 12:00:38 Wire Position: X0 = 1.0000 mm, Y0 = 0.0000 mm 12:00:41 Integrated Voltage Measurement... 12:00:50 Voltage Offset Correction: v_off_ini = 446.874 uV, v_off_fin = 430.943 uV 12:00:50 Integrated Voltage Measurement dXYus = -1.0001 mm, dXYds = 0.0000 mm dXY = -1.0001 mm VT = -113.034 uVs 12:00:50 Integrated Voltage Measurement... 12:00:59 Voltage Offset Correction: v_off_ini = 436.266 uV, v_off_fin = 441.976 uV 12:00:59 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 112.798 uVs 12:00:59 Integrated Voltage Measurement... 12:01:08 Voltage Offset Correction: v_off_ini = 427.131 uV, v_off_fin = 423.165 uV 12:01:08 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -103.925 uVs 12:01:08 Integrated Voltage Measurement... 12:01:17 Voltage Offset Correction: v_off_ini = 431.027 uV, v_off_fin = 432.113 uV 12:01:17 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 107.788 uVs 12:01:17 Integrated Voltage Measurement... 12:01:26 Voltage Offset Correction: v_off_ini = 424.082 uV, v_off_fin = 426.474 uV 12:01:26 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -92.440 uVs 12:01:26 Integrated Voltage Measurement... 12:01:35 Voltage Offset Correction: v_off_ini = 425.512 uV, v_off_fin = 439.199 uV 12:01:35 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 117.417 uVs 12:01:37 I2X Average and RMS of Measurements VT = -107.900 +- 3.567 uVs I2X = 9.927 +- 0.328 uTm^2 12:01:42 Wire Position: X0 = 1.0000 mm, Y0 = 0.0000 mm 12:01:45 Integrated Voltage Measurement... 12:01:54 Voltage Offset Correction: v_off_ini = 440.937 uV, v_off_fin = 443.281 uV 12:01:54 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -576.044 uVs 12:01:54 Integrated Voltage Measurement... 12:02:03 Voltage Offset Correction: v_off_ini = 443.847 uV, v_off_fin = 431.099 uV 12:02:03 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 553.702 uVs 12:02:03 Integrated Voltage Measurement... 12:02:12 Voltage Offset Correction: v_off_ini = 439.776 uV, v_off_fin = 439.931 uV 12:02:12 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -564.440 uVs 12:02:12 Integrated Voltage Measurement... 12:02:21 Voltage Offset Correction: v_off_ini = 454.092 uV, v_off_fin = 434.304 uV 12:02:21 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 537.548 uVs 12:02:21 Integrated Voltage Measurement... 12:02:30 Voltage Offset Correction: v_off_ini = 420.682 uV, v_off_fin = 424.539 uV 12:02:30 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -572.168 uVs 12:02:30 Integrated Voltage Measurement... 12:02:39 Voltage Offset Correction: v_off_ini = 427.026 uV, v_off_fin = 427.174 uV 12:02:39 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 578.731 uVs 12:02:42 I2Y Average and RMS of Measurements VT = -563.772 +- 10.014 uVs I2Y = -51.867 +- 0.921 uTm^2 12:02:46 Wire Position: X0 = 2.0000 mm, Y0 = 0.0000 mm 12:02:51 Integrated Voltage Measurement... 12:03:00 Voltage Offset Correction: v_off_ini = 442.669 uV, v_off_fin = 424.414 uV 12:03:00 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.4998 mm dXY = -0.4999 mm VT = -107.282 uVs 12:03:00 Integrated Voltage Measurement... 12:03:09 Voltage Offset Correction: v_off_ini = 419.163 uV, v_off_fin = 431.144 uV 12:03:09 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = 104.542 uVs 12:03:09 Integrated Voltage Measurement... 12:03:18 Voltage Offset Correction: v_off_ini = 434.187 uV, v_off_fin = 431.364 uV 12:03:18 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -97.764 uVs 12:03:18 Integrated Voltage Measurement... 12:03:27 Voltage Offset Correction: v_off_ini = 440.873 uV, v_off_fin = 432.924 uV 12:03:27 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 105.357 uVs 12:03:27 Integrated Voltage Measurement... 12:03:36 Voltage Offset Correction: v_off_ini = 431.970 uV, v_off_fin = 422.479 uV 12:03:36 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -103.209 uVs 12:03:36 Integrated Voltage Measurement... 12:03:45 Voltage Offset Correction: v_off_ini = 428.066 uV, v_off_fin = 439.807 uV 12:03:45 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 86.366 uVs 12:03:49 I1X Average and RMS of Measurements VT = -100.753 +- 4.577 uVs I1X = 20.153 +- 0.917 uTm 12:03:51 Wire Position: X0 = 2.0000 mm, Y0 = 0.0000 mm 12:03:56 Integrated Voltage Measurement... 12:04:05 Voltage Offset Correction: v_off_ini = 462.602 uV, v_off_fin = 439.923 uV 12:04:05 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -319.089 uVs 12:04:05 Integrated Voltage Measurement... 12:04:14 Voltage Offset Correction: v_off_ini = 448.621 uV, v_off_fin = 463.588 uV 12:04:14 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 287.152 uVs 12:04:14 Integrated Voltage Measurement... 12:04:23 Voltage Offset Correction: v_off_ini = 450.049 uV, v_off_fin = 429.761 uV 12:04:23 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -311.102 uVs 12:04:23 Integrated Voltage Measurement... 12:04:32 Voltage Offset Correction: v_off_ini = 447.956 uV, v_off_fin = 449.104 uV 12:04:32 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 276.175 uVs 12:04:32 Integrated Voltage Measurement... 12:04:41 Voltage Offset Correction: v_off_ini = 454.253 uV, v_off_fin = 438.166 uV 12:04:41 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -288.299 uVs 12:04:41 Integrated Voltage Measurement... 12:04:50 Voltage Offset Correction: v_off_ini = 444.100 uV, v_off_fin = 457.991 uV 12:04:50 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 287.170 uVs 12:04:54 I1Y Average and RMS of Measurements VT = -294.831 +- 6.338 uVs I1Y = -58.966 +- 1.268 uTm 12:04:59 Wire Position: X0 = 2.0000 mm, Y0 = 0.0000 mm 12:05:02 Integrated Voltage Measurement... 12:05:11 Voltage Offset Correction: v_off_ini = 429.338 uV, v_off_fin = 444.146 uV 12:05:11 Integrated Voltage Measurement dXYus = -1.0002 mm, dXYds = 0.0000 mm dXY = -1.0002 mm VT = -127.243 uVs 12:05:11 Integrated Voltage Measurement... 12:05:20 Voltage Offset Correction: v_off_ini = 460.280 uV, v_off_fin = 461.992 uV 12:05:20 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 111.993 uVs 12:05:20 Integrated Voltage Measurement... 12:05:29 Voltage Offset Correction: v_off_ini = 446.013 uV, v_off_fin = 438.637 uV 12:05:29 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -120.216 uVs 12:05:29 Integrated Voltage Measurement... 12:05:38 Voltage Offset Correction: v_off_ini = 447.726 uV, v_off_fin = 456.652 uV 12:05:38 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 123.779 uVs 12:05:38 Integrated Voltage Measurement... 12:05:47 Voltage Offset Correction: v_off_ini = 464.137 uV, v_off_fin = 449.307 uV 12:05:47 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -135.876 uVs 12:05:47 Integrated Voltage Measurement... 12:05:56 Voltage Offset Correction: v_off_ini = 458.077 uV, v_off_fin = 457.411 uV 12:05:56 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 132.784 uVs 12:05:58 I2X Average and RMS of Measurements VT = -125.315 +- 6.448 uVs I2X = 11.529 +- 0.593 uTm^2 12:06:03 Wire Position: X0 = 2.0000 mm, Y0 = 0.0000 mm 12:06:06 Integrated Voltage Measurement... 12:06:15 Voltage Offset Correction: v_off_ini = 458.187 uV, v_off_fin = 460.114 uV 12:06:15 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -575.450 uVs 12:06:15 Integrated Voltage Measurement... 12:06:24 Voltage Offset Correction: v_off_ini = 460.493 uV, v_off_fin = 458.877 uV 12:06:24 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 560.257 uVs 12:06:24 Integrated Voltage Measurement... 12:06:33 Voltage Offset Correction: v_off_ini = 455.315 uV, v_off_fin = 441.541 uV 12:06:33 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -555.629 uVs 12:06:33 Integrated Voltage Measurement... 12:06:42 Voltage Offset Correction: v_off_ini = 455.872 uV, v_off_fin = 460.341 uV 12:06:42 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 566.693 uVs 12:06:42 Integrated Voltage Measurement... 12:06:51 Voltage Offset Correction: v_off_ini = 448.285 uV, v_off_fin = 459.293 uV 12:06:51 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -560.448 uVs 12:06:51 Integrated Voltage Measurement... 12:07:00 Voltage Offset Correction: v_off_ini = 459.843 uV, v_off_fin = 447.595 uV 12:07:00 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 576.627 uVs 12:07:03 I2Y Average and RMS of Measurements VT = -565.851 +- 3.328 uVs I2Y = -52.058 +- 0.306 uTm^2 12:07:10 Wire Position: X0 = 0.0000 mm, Y0 = -1.0000 mm 12:07:14 Integrated Voltage Measurement... 12:07:23 Voltage Offset Correction: v_off_ini = 477.647 uV, v_off_fin = 470.643 uV 12:07:23 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.4998 mm dXY = -0.4999 mm VT = -60.951 uVs 12:07:24 Integrated Voltage Measurement... 12:07:32 Voltage Offset Correction: v_off_ini = 468.104 uV, v_off_fin = 467.736 uV 12:07:33 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 79.058 uVs 12:07:33 Integrated Voltage Measurement... 12:07:41 Voltage Offset Correction: v_off_ini = 470.784 uV, v_off_fin = 462.606 uV 12:07:42 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -74.657 uVs 12:07:42 Integrated Voltage Measurement... 12:07:51 Voltage Offset Correction: v_off_ini = 468.200 uV, v_off_fin = 457.553 uV 12:07:51 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 43.375 uVs 12:07:51 Integrated Voltage Measurement... 12:08:00 Voltage Offset Correction: v_off_ini = 464.190 uV, v_off_fin = 470.561 uV 12:08:00 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -60.408 uVs 12:08:00 Integrated Voltage Measurement... 12:08:09 Voltage Offset Correction: v_off_ini = 458.864 uV, v_off_fin = 469.363 uV 12:08:09 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.4999 mm dXY = 0.4999 mm VT = 55.652 uVs 12:08:13 I1X Average and RMS of Measurements VT = -62.350 +- 5.427 uVs I1X = 12.471 +- 1.086 uTm 12:08:15 Wire Position: X0 = 0.0000 mm, Y0 = -1.0000 mm 12:08:20 Integrated Voltage Measurement... 12:08:29 Voltage Offset Correction: v_off_ini = 480.323 uV, v_off_fin = 464.592 uV 12:08:29 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -274.862 uVs 12:08:29 Integrated Voltage Measurement... 12:08:37 Voltage Offset Correction: v_off_ini = 482.704 uV, v_off_fin = 485.113 uV 12:08:38 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 263.589 uVs 12:08:38 Integrated Voltage Measurement... 12:08:47 Voltage Offset Correction: v_off_ini = 471.901 uV, v_off_fin = 461.802 uV 12:08:47 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -264.505 uVs 12:08:47 Integrated Voltage Measurement... 12:08:56 Voltage Offset Correction: v_off_ini = 477.068 uV, v_off_fin = 476.683 uV 12:08:56 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 266.142 uVs 12:08:56 Integrated Voltage Measurement... 12:09:04 Voltage Offset Correction: v_off_ini = 473.003 uV, v_off_fin = 465.660 uV 12:09:05 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -268.503 uVs 12:09:05 Integrated Voltage Measurement... 12:09:15 Voltage Offset Correction: v_off_ini = 477.892 uV, v_off_fin = 489.533 uV 12:09:15 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 281.024 uVs 12:09:20 I1Y Average and RMS of Measurements VT = -269.771 +- 3.873 uVs I1Y = -53.954 +- 0.775 uTm 12:09:22 Wire Position: X0 = 0.0000 mm, Y0 = -1.0000 mm 12:09:25 Integrated Voltage Measurement... 12:09:34 Voltage Offset Correction: v_off_ini = 463.053 uV, v_off_fin = 451.500 uV 12:09:34 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -62.609 uVs 12:09:34 Integrated Voltage Measurement... 12:09:43 Voltage Offset Correction: v_off_ini = 473.496 uV, v_off_fin = 488.270 uV 12:09:43 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 94.151 uVs 12:09:43 Integrated Voltage Measurement... 12:09:52 Voltage Offset Correction: v_off_ini = 479.482 uV, v_off_fin = 463.888 uV 12:09:52 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -81.677 uVs 12:09:52 Integrated Voltage Measurement... 12:10:01 Voltage Offset Correction: v_off_ini = 481.038 uV, v_off_fin = 488.001 uV 12:10:01 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 64.453 uVs 12:10:01 Integrated Voltage Measurement... 12:10:10 Voltage Offset Correction: v_off_ini = 465.088 uV, v_off_fin = 470.435 uV 12:10:10 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -71.911 uVs 12:10:10 Integrated Voltage Measurement... 12:10:19 Voltage Offset Correction: v_off_ini = 486.146 uV, v_off_fin = 489.106 uV 12:10:19 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 60.710 uVs 12:10:21 I2X Average and RMS of Measurements VT = -72.585 +- 4.939 uVs I2X = 6.678 +- 0.454 uTm^2 12:10:25 Wire Position: X0 = 0.0000 mm, Y0 = -1.0000 mm 14:17:22 Integrated Voltage Measurement... 14:17:31 Voltage Offset Correction: v_off_ini = 620.982 uV, v_off_fin = 621.418 uV 14:17:31 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -538.709 uVs 14:17:31 Integrated Voltage Measurement... 14:17:40 Voltage Offset Correction: v_off_ini = 615.802 uV, v_off_fin = 613.234 uV 14:17:40 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 533.870 uVs 14:17:40 Integrated Voltage Measurement... 14:17:49 Voltage Offset Correction: v_off_ini = 609.542 uV, v_off_fin = 598.840 uV 14:17:49 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -539.247 uVs 14:17:49 Integrated Voltage Measurement... 14:17:58 Voltage Offset Correction: v_off_ini = 608.037 uV, v_off_fin = 612.447 uV 14:17:58 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 538.702 uVs 14:17:58 Integrated Voltage Measurement... 14:18:07 Voltage Offset Correction: v_off_ini = 616.844 uV, v_off_fin = 618.148 uV 14:18:07 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -554.797 uVs 14:18:07 Integrated Voltage Measurement... 14:18:16 Voltage Offset Correction: v_off_ini = 624.934 uV, v_off_fin = 612.786 uV 14:18:16 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 540.386 uVs 14:18:18 I2Y Average and RMS of Measurements VT = -540.952 +- 4.821 uVs I2Y = -49.768 +- 0.444 uTm^2 14:18:22 Wire Position: X0 = 0.0000 mm, Y0 = -0.5000 mm 14:18:27 Integrated Voltage Measurement... 14:18:36 Voltage Offset Correction: v_off_ini = 613.812 uV, v_off_fin = 617.048 uV 14:18:36 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.4998 mm dXY = -0.4999 mm VT = -68.025 uVs 14:18:36 Integrated Voltage Measurement... 14:18:45 Voltage Offset Correction: v_off_ini = 627.711 uV, v_off_fin = 635.940 uV 14:18:45 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 60.066 uVs 14:18:45 Integrated Voltage Measurement... 14:18:54 Voltage Offset Correction: v_off_ini = 618.278 uV, v_off_fin = 617.770 uV 14:18:54 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -65.912 uVs 14:18:54 Integrated Voltage Measurement... 14:19:03 Voltage Offset Correction: v_off_ini = 624.815 uV, v_off_fin = 618.503 uV 14:19:03 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 68.464 uVs 14:19:03 Integrated Voltage Measurement... 14:19:12 Voltage Offset Correction: v_off_ini = 616.595 uV, v_off_fin = 615.746 uV 14:19:12 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -70.827 uVs 14:19:12 Integrated Voltage Measurement... 14:19:21 Voltage Offset Correction: v_off_ini = 618.134 uV, v_off_fin = 624.473 uV 14:19:21 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 55.217 uVs 14:19:25 I1X Average and RMS of Measurements VT = -64.752 +- 1.773 uVs I1X = 12.951 +- 0.354 uTm 14:19:27 Wire Position: X0 = 0.0000 mm, Y0 = -0.5000 mm 14:19:32 Integrated Voltage Measurement... 14:19:41 Voltage Offset Correction: v_off_ini = 605.625 uV, v_off_fin = 602.440 uV 14:19:41 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = -283.707 uVs 14:19:41 Integrated Voltage Measurement... 14:19:50 Voltage Offset Correction: v_off_ini = 618.520 uV, v_off_fin = 620.512 uV 14:19:50 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 283.136 uVs 14:19:50 Integrated Voltage Measurement... 14:19:59 Voltage Offset Correction: v_off_ini = 615.305 uV, v_off_fin = 608.160 uV 14:19:59 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -263.159 uVs 14:19:59 Integrated Voltage Measurement... 14:20:08 Voltage Offset Correction: v_off_ini = 616.427 uV, v_off_fin = 629.051 uV 14:20:08 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 269.900 uVs 14:20:08 Integrated Voltage Measurement... 14:20:17 Voltage Offset Correction: v_off_ini = 598.633 uV, v_off_fin = 599.076 uV 14:20:17 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -271.739 uVs 14:20:17 Integrated Voltage Measurement... 14:20:26 Voltage Offset Correction: v_off_ini = 602.838 uV, v_off_fin = 614.345 uV 14:20:26 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 273.122 uVs 14:20:30 I1Y Average and RMS of Measurements VT = -274.127 +- 7.000 uVs I1Y = -54.826 +- 1.401 uTm 14:20:32 Wire Position: X0 = 0.0000 mm, Y0 = -0.5000 mm 14:20:35 Integrated Voltage Measurement... 14:20:44 Voltage Offset Correction: v_off_ini = 602.390 uV, v_off_fin = 603.999 uV 14:20:44 Integrated Voltage Measurement dXYus = -1.0001 mm, dXYds = 0.0000 mm dXY = -1.0001 mm VT = -88.626 uVs 14:20:44 Integrated Voltage Measurement... 14:20:53 Voltage Offset Correction: v_off_ini = 619.024 uV, v_off_fin = 613.046 uV 14:20:53 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 80.756 uVs 14:20:53 Integrated Voltage Measurement... 14:21:02 Voltage Offset Correction: v_off_ini = 606.369 uV, v_off_fin = 602.356 uV 14:21:02 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -85.250 uVs 14:21:02 Integrated Voltage Measurement... 14:21:11 Voltage Offset Correction: v_off_ini = 602.228 uV, v_off_fin = 621.821 uV 14:21:11 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 87.112 uVs 14:21:11 Integrated Voltage Measurement... 14:21:20 Voltage Offset Correction: v_off_ini = 608.690 uV, v_off_fin = 598.422 uV 14:21:20 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -76.930 uVs 14:21:20 Integrated Voltage Measurement... 14:21:29 Voltage Offset Correction: v_off_ini = 602.567 uV, v_off_fin = 602.156 uV 14:21:29 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 70.581 uVs 14:21:31 I2X Average and RMS of Measurements VT = -81.543 +- 5.540 uVs I2X = 7.502 +- 0.509 uTm^2 14:21:34 Wire Position: X0 = 0.0000 mm, Y0 = -0.5000 mm 14:21:36 Integrated Voltage Measurement... 14:21:45 Voltage Offset Correction: v_off_ini = 599.117 uV, v_off_fin = 604.178 uV 14:21:45 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -544.504 uVs 14:21:46 Integrated Voltage Measurement... 14:21:54 Voltage Offset Correction: v_off_ini = 601.807 uV, v_off_fin = 609.852 uV 14:21:54 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 555.156 uVs 14:21:55 Integrated Voltage Measurement... 14:22:03 Voltage Offset Correction: v_off_ini = 610.618 uV, v_off_fin = 595.150 uV 14:22:04 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -545.524 uVs 14:22:04 Integrated Voltage Measurement... 14:22:13 Voltage Offset Correction: v_off_ini = 609.598 uV, v_off_fin = 617.958 uV 14:22:13 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 563.361 uVs 14:22:13 Integrated Voltage Measurement... 14:22:22 Voltage Offset Correction: v_off_ini = 610.259 uV, v_off_fin = 603.959 uV 14:22:22 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -539.739 uVs 14:22:22 Integrated Voltage Measurement... 14:22:31 Voltage Offset Correction: v_off_ini = 611.036 uV, v_off_fin = 618.467 uV 14:22:31 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 548.317 uVs 14:22:33 I2Y Average and RMS of Measurements VT = -549.434 +- 4.261 uVs I2Y = -50.548 +- 0.392 uTm^2 14:22:37 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 14:22:41 Integrated Voltage Measurement... 14:22:50 Voltage Offset Correction: v_off_ini = 612.579 uV, v_off_fin = 610.988 uV 14:22:50 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -77.741 uVs 14:22:50 Integrated Voltage Measurement... 14:22:59 Voltage Offset Correction: v_off_ini = 613.051 uV, v_off_fin = 603.179 uV 14:22:59 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 92.428 uVs 14:22:59 Integrated Voltage Measurement... 14:23:08 Voltage Offset Correction: v_off_ini = 611.811 uV, v_off_fin = 614.912 uV 14:23:08 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -53.573 uVs 14:23:08 Integrated Voltage Measurement... 14:23:17 Voltage Offset Correction: v_off_ini = 622.552 uV, v_off_fin = 624.838 uV 14:23:17 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 72.970 uVs 14:23:17 Integrated Voltage Measurement... 14:23:26 Voltage Offset Correction: v_off_ini = 606.089 uV, v_off_fin = 612.945 uV 14:23:26 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -75.585 uVs 14:23:26 Integrated Voltage Measurement... 14:23:35 Voltage Offset Correction: v_off_ini = 620.080 uV, v_off_fin = 606.244 uV 14:23:35 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 69.970 uVs 14:23:39 I1X Average and RMS of Measurements VT = -73.711 +- 8.929 uVs I1X = 14.743 +- 1.786 uTm 14:23:42 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 14:23:46 Integrated Voltage Measurement... 14:23:55 Voltage Offset Correction: v_off_ini = 611.615 uV, v_off_fin = 596.717 uV 14:23:55 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -280.685 uVs 14:23:55 Integrated Voltage Measurement... 14:24:04 Voltage Offset Correction: v_off_ini = 593.981 uV, v_off_fin = 624.979 uV 14:24:04 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 296.792 uVs 14:24:04 Integrated Voltage Measurement... 14:24:13 Voltage Offset Correction: v_off_ini = 617.899 uV, v_off_fin = 599.403 uV 14:24:13 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -299.994 uVs 14:24:13 Integrated Voltage Measurement... 14:24:22 Voltage Offset Correction: v_off_ini = 588.805 uV, v_off_fin = 591.095 uV 14:24:22 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 288.336 uVs 14:24:22 Integrated Voltage Measurement... 14:24:31 Voltage Offset Correction: v_off_ini = 566.166 uV, v_off_fin = 552.720 uV 14:24:31 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -300.773 uVs 14:24:31 Integrated Voltage Measurement... 14:24:40 Voltage Offset Correction: v_off_ini = 585.630 uV, v_off_fin = 606.211 uV 14:24:40 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 282.307 uVs 14:24:44 I1Y Average and RMS of Measurements VT = -291.481 +- 2.216 uVs I1Y = -58.296 +- 0.443 uTm 14:24:47 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 14:24:49 Integrated Voltage Measurement... 14:24:58 Voltage Offset Correction: v_off_ini = 594.763 uV, v_off_fin = 597.100 uV 14:24:58 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -66.966 uVs 14:24:58 Integrated Voltage Measurement... 14:25:07 Voltage Offset Correction: v_off_ini = 588.260 uV, v_off_fin = 597.657 uV 14:25:07 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 89.931 uVs 14:25:07 Integrated Voltage Measurement... 14:25:16 Voltage Offset Correction: v_off_ini = 590.814 uV, v_off_fin = 580.026 uV 14:25:16 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -87.077 uVs 14:25:16 Integrated Voltage Measurement... 14:25:25 Voltage Offset Correction: v_off_ini = 604.568 uV, v_off_fin = 602.513 uV 14:25:25 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 79.198 uVs 14:25:25 Integrated Voltage Measurement... 14:25:34 Voltage Offset Correction: v_off_ini = 587.193 uV, v_off_fin = 594.569 uV 14:25:34 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -92.161 uVs 14:25:34 Integrated Voltage Measurement... 14:25:43 Voltage Offset Correction: v_off_ini = 601.209 uV, v_off_fin = 600.913 uV 14:25:43 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 86.309 uVs 14:25:46 I2X Average and RMS of Measurements VT = -83.607 +- 4.416 uVs I2X = 7.692 +- 0.406 uTm^2 14:25:48 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 14:25:51 Integrated Voltage Measurement... 14:26:00 Voltage Offset Correction: v_off_ini = 592.223 uV, v_off_fin = 596.859 uV 14:26:00 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -544.221 uVs 14:26:00 Integrated Voltage Measurement... 14:26:09 Voltage Offset Correction: v_off_ini = 590.951 uV, v_off_fin = 596.788 uV 14:26:09 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 562.039 uVs 14:26:09 Integrated Voltage Measurement... 14:26:18 Voltage Offset Correction: v_off_ini = 597.189 uV, v_off_fin = 600.767 uV 14:26:18 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -567.526 uVs 14:26:18 Integrated Voltage Measurement... 14:26:27 Voltage Offset Correction: v_off_ini = 595.816 uV, v_off_fin = 602.428 uV 14:26:27 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 580.491 uVs 14:26:27 Integrated Voltage Measurement... 14:26:36 Voltage Offset Correction: v_off_ini = 587.483 uV, v_off_fin = 582.325 uV 14:26:36 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -542.297 uVs 14:26:36 Integrated Voltage Measurement... 14:26:45 Voltage Offset Correction: v_off_ini = 581.363 uV, v_off_fin = 599.131 uV 14:26:45 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 565.132 uVs 14:26:47 I2Y Average and RMS of Measurements VT = -560.284 +- 9.707 uVs I2Y = -51.546 +- 0.893 uTm^2 14:26:51 Wire Position: X0 = 0.0000 mm, Y0 = 0.5000 mm 14:26:56 Integrated Voltage Measurement... 14:27:05 Voltage Offset Correction: v_off_ini = 603.644 uV, v_off_fin = 592.521 uV 14:27:05 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -56.189 uVs 14:27:05 Integrated Voltage Measurement... 14:27:14 Voltage Offset Correction: v_off_ini = 587.076 uV, v_off_fin = 586.881 uV 14:27:14 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.4999 mm dXY = 0.4999 mm VT = 41.554 uVs 14:27:14 Integrated Voltage Measurement... 14:27:23 Voltage Offset Correction: v_off_ini = 584.330 uV, v_off_fin = 588.467 uV 14:27:23 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -54.250 uVs 14:27:23 Integrated Voltage Measurement... 14:27:32 Voltage Offset Correction: v_off_ini = 600.131 uV, v_off_fin = 588.169 uV 14:27:32 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = 47.030 uVs 14:27:32 Integrated Voltage Measurement... 14:27:41 Voltage Offset Correction: v_off_ini = 575.454 uV, v_off_fin = 577.813 uV 14:27:41 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -56.479 uVs 14:27:41 Integrated Voltage Measurement... 14:27:50 Voltage Offset Correction: v_off_ini = 594.329 uV, v_off_fin = 602.176 uV 14:27:50 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 56.690 uVs 14:27:54 I1X Average and RMS of Measurements VT = -52.032 +- 3.299 uVs I1X = 10.407 +- 0.659 uTm 14:27:56 Wire Position: X0 = 0.0000 mm, Y0 = 0.5000 mm 14:28:01 Integrated Voltage Measurement... 14:28:10 Voltage Offset Correction: v_off_ini = 588.097 uV, v_off_fin = 578.352 uV 14:28:10 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -308.863 uVs 14:28:10 Integrated Voltage Measurement... 14:28:19 Voltage Offset Correction: v_off_ini = 590.438 uV, v_off_fin = 600.443 uV 14:28:19 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 291.389 uVs 14:28:19 Integrated Voltage Measurement... 14:28:28 Voltage Offset Correction: v_off_ini = 596.963 uV, v_off_fin = 577.419 uV 14:28:28 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -294.638 uVs 14:28:28 Integrated Voltage Measurement... 14:28:37 Voltage Offset Correction: v_off_ini = 595.330 uV, v_off_fin = 592.684 uV 14:28:37 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 304.764 uVs 14:28:37 Integrated Voltage Measurement... 14:28:46 Voltage Offset Correction: v_off_ini = 583.091 uV, v_off_fin = 575.929 uV 14:28:46 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -294.754 uVs 14:28:46 Integrated Voltage Measurement... 14:28:55 Voltage Offset Correction: v_off_ini = 592.681 uV, v_off_fin = 606.101 uV 14:28:55 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 285.461 uVs 14:28:59 I1Y Average and RMS of Measurements VT = -296.645 +- 4.626 uVs I1Y = -59.323 +- 0.925 uTm 14:29:01 Wire Position: X0 = 0.0000 mm, Y0 = 0.5000 mm 14:29:04 Integrated Voltage Measurement... 14:29:13 Voltage Offset Correction: v_off_ini = 577.673 uV, v_off_fin = 589.332 uV 14:29:13 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -64.363 uVs 14:29:13 Integrated Voltage Measurement... 14:29:22 Voltage Offset Correction: v_off_ini = 591.220 uV, v_off_fin = 606.108 uV 14:29:22 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 80.124 uVs 14:29:22 Integrated Voltage Measurement... 14:29:31 Voltage Offset Correction: v_off_ini = 593.376 uV, v_off_fin = 596.083 uV 14:29:31 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -77.072 uVs 14:29:31 Integrated Voltage Measurement... 14:29:40 Voltage Offset Correction: v_off_ini = 595.352 uV, v_off_fin = 617.932 uV 14:29:40 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 91.167 uVs 14:29:40 Integrated Voltage Measurement... 14:29:49 Voltage Offset Correction: v_off_ini = 622.003 uV, v_off_fin = 605.820 uV 14:29:49 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -83.893 uVs 14:29:49 Integrated Voltage Measurement... 14:29:58 Voltage Offset Correction: v_off_ini = 609.938 uV, v_off_fin = 611.778 uV 14:29:58 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 93.069 uVs 14:30:00 I2X Average and RMS of Measurements VT = -81.615 +- 6.862 uVs I2X = 7.509 +- 0.631 uTm^2 14:30:03 Wire Position: X0 = 0.0000 mm, Y0 = 0.5000 mm 14:30:05 Integrated Voltage Measurement... 14:30:14 Voltage Offset Correction: v_off_ini = 602.545 uV, v_off_fin = 611.397 uV 14:30:14 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -570.671 uVs 14:30:14 Integrated Voltage Measurement... 14:30:23 Voltage Offset Correction: v_off_ini = 605.996 uV, v_off_fin = 605.255 uV 14:30:23 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 576.465 uVs 14:30:23 Integrated Voltage Measurement... 14:30:32 Voltage Offset Correction: v_off_ini = 596.787 uV, v_off_fin = 605.304 uV 14:30:32 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -560.455 uVs 14:30:33 Integrated Voltage Measurement... 14:30:41 Voltage Offset Correction: v_off_ini = 604.698 uV, v_off_fin = 628.884 uV 14:30:42 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 577.560 uVs 14:30:42 Integrated Voltage Measurement... 14:30:51 Voltage Offset Correction: v_off_ini = 616.500 uV, v_off_fin = 612.080 uV 14:30:51 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -559.776 uVs 14:30:51 Integrated Voltage Measurement... 14:31:00 Voltage Offset Correction: v_off_ini = 602.992 uV, v_off_fin = 602.120 uV 14:31:00 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 587.179 uVs 14:31:02 I2Y Average and RMS of Measurements VT = -572.018 +- 2.129 uVs I2Y = -52.626 +- 0.196 uTm^2