SLAC Magnetic Measurements Date: 07-23-2026 Time: 14:03:37 Project: LCLS-II-HE Magnet Type: Phase Shifter Magnet Name: HE_SXPS_16336 Dataset Number: 1 Run Purpose: Tuning Run Type: Run Number: 3 Measurement Device: Stretched Wire Operator: YL Comment: 14:03:45 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 14:03:47 Motion Parameters, X Controller, Axis 1 acc = 0.001000 m/s/s dcl = 0.001000 m/s/s vel = 0.001000 m/s pos = 0.000000 m kp = 188.600000 ki = 820.000000 kd = 491.400000 db = 0.000000 14:03:49 Motion Parameters, X Controller, Axis 2 acc = 0.001000 m/s/s dcl = 0.001000 m/s/s vel = 0.001000 m/s pos = 0.000000 m kp = 188.600000 ki = 820.000000 kd = 491.400000 db = 0.000000 14:03:50 Motion Parameters, Y Controller, Axis 1 acc = 0.001000 m/s/s dcl = 0.001000 m/s/s vel = 0.001000 m/s pos = 0.000000 m kp = 188.600000 ki = 820.000000 kd = 491.400000 db = 0.000000 14:03:52 Motion Parameters, Y Controller, Axis 2 acc = 0.001000 m/s/s dcl = 0.001000 m/s/s vel = 0.001000 m/s pos = 0.000000 m kp = 188.600000 ki = 820.000000 kd = 491.400000 db = 0.000000 14:04:24 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 14:04:28 Integrated Voltage Measurement... 14:04:37 Voltage Offset Correction: v_off_ini = 2302.048 uV, v_off_fin = 2290.870 uV 14:04:37 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -68.691 uVs 14:04:38 Integrated Voltage Measurement... 14:04:46 Voltage Offset Correction: v_off_ini = 2293.791 uV, v_off_fin = 2287.153 uV 14:04:46 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 82.855 uVs 14:04:47 Integrated Voltage Measurement... 14:04:55 Voltage Offset Correction: v_off_ini = 2316.898 uV, v_off_fin = 2286.499 uV 14:04:55 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -76.442 uVs 14:04:56 Integrated Voltage Measurement... 14:05:04 Voltage Offset Correction: v_off_ini = 2302.221 uV, v_off_fin = 2315.001 uV 14:05:04 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 62.171 uVs 14:05:05 Integrated Voltage Measurement... 14:05:13 Voltage Offset Correction: v_off_ini = 2327.439 uV, v_off_fin = 2311.902 uV 14:05:14 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -102.072 uVs 14:05:14 Integrated Voltage Measurement... 14:05:22 Voltage Offset Correction: v_off_ini = 2283.291 uV, v_off_fin = 2297.349 uV 14:05:23 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 101.638 uVs 14:05:27 I1X Average and RMS of Measurements VT = -82.312 +- 14.069 uVs I1X = 16.462 +- 2.814 uTm 14:05:29 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 14:05:33 Integrated Voltage Measurement... 14:05:42 Voltage Offset Correction: v_off_ini = 2307.171 uV, v_off_fin = 2307.852 uV 14:05:42 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -254.208 uVs 14:05:42 Integrated Voltage Measurement... 14:05:51 Voltage Offset Correction: v_off_ini = 2304.446 uV, v_off_fin = 2320.115 uV 14:05:51 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 262.961 uVs 14:05:51 Integrated Voltage Measurement... 14:06:00 Voltage Offset Correction: v_off_ini = 2294.735 uV, v_off_fin = 2305.275 uV 14:06:00 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -289.216 uVs 14:06:00 Integrated Voltage Measurement... 14:06:09 Voltage Offset Correction: v_off_ini = 2326.108 uV, v_off_fin = 2329.282 uV 14:06:09 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 262.864 uVs 14:06:09 Integrated Voltage Measurement... 14:06:18 Voltage Offset Correction: v_off_ini = 2343.093 uV, v_off_fin = 2314.795 uV 14:06:18 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -257.485 uVs 14:06:18 Integrated Voltage Measurement... 14:06:27 Voltage Offset Correction: v_off_ini = 2321.606 uV, v_off_fin = 2331.559 uV 14:06:27 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 287.571 uVs 14:06:32 I1Y Average and RMS of Measurements VT = -269.051 +- 7.539 uVs I1Y = -53.805 +- 1.508 uTm 14:06:37 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 14:06:39 Integrated Voltage Measurement... 14:06:48 Voltage Offset Correction: v_off_ini = 2337.157 uV, v_off_fin = 2370.381 uV 14:06:48 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -50.744 uVs 14:06:48 Integrated Voltage Measurement... 14:06:57 Voltage Offset Correction: v_off_ini = 2372.231 uV, v_off_fin = 2384.029 uV 14:06:57 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 45.596 uVs 14:06:57 Integrated Voltage Measurement... 14:07:06 Voltage Offset Correction: v_off_ini = 2375.395 uV, v_off_fin = 2366.171 uV 14:07:06 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -62.323 uVs 14:07:06 Integrated Voltage Measurement... 14:07:15 Voltage Offset Correction: v_off_ini = 2369.293 uV, v_off_fin = 2364.938 uV 14:07:15 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 69.283 uVs 14:07:15 Integrated Voltage Measurement... 14:07:24 Voltage Offset Correction: v_off_ini = 2362.619 uV, v_off_fin = 2343.611 uV 14:07:24 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -77.147 uVs 14:07:24 Integrated Voltage Measurement... 14:07:33 Voltage Offset Correction: v_off_ini = 2335.015 uV, v_off_fin = 2323.617 uV 14:07:33 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 56.996 uVs 14:07:36 I2X Average and RMS of Measurements VT = -60.348 +- 8.627 uVs I2X = 5.552 +- 0.794 uTm^2 14:07:41 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 14:07:43 Integrated Voltage Measurement... 14:07:52 Voltage Offset Correction: v_off_ini = 2344.206 uV, v_off_fin = 2354.426 uV 14:07:52 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -416.282 uVs 14:07:52 Integrated Voltage Measurement... 14:08:01 Voltage Offset Correction: v_off_ini = 2347.701 uV, v_off_fin = 2329.875 uV 14:08:01 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 420.974 uVs 14:08:01 Integrated Voltage Measurement... 14:08:10 Voltage Offset Correction: v_off_ini = 2344.780 uV, v_off_fin = 2326.914 uV 14:08:10 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -410.321 uVs 14:08:10 Integrated Voltage Measurement... 14:08:19 Voltage Offset Correction: v_off_ini = 2322.147 uV, v_off_fin = 2314.292 uV 14:08:19 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 397.918 uVs 14:08:20 Integrated Voltage Measurement... 14:08:28 Voltage Offset Correction: v_off_ini = 2288.406 uV, v_off_fin = 2298.663 uV 14:08:29 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -392.461 uVs 14:08:29 Integrated Voltage Measurement... 14:08:38 Voltage Offset Correction: v_off_ini = 2297.165 uV, v_off_fin = 2301.930 uV 14:08:38 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 406.387 uVs 14:08:40 I2Y Average and RMS of Measurements VT = -407.391 +- 8.174 uVs I2Y = -37.480 +- 0.752 uTm^2 14:08:51 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 14:08:55 Integrated Voltage Measurement... 14:09:04 Voltage Offset Correction: v_off_ini = 2325.801 uV, v_off_fin = 2302.263 uV 14:09:04 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -101.064 uVs 14:09:04 Integrated Voltage Measurement... 14:09:13 Voltage Offset Correction: v_off_ini = 2346.929 uV, v_off_fin = 2333.347 uV 14:09:13 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 75.407 uVs 14:09:13 Integrated Voltage Measurement... 14:09:22 Voltage Offset Correction: v_off_ini = 2342.873 uV, v_off_fin = 2355.848 uV 14:09:22 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -58.773 uVs 14:09:22 Integrated Voltage Measurement... 14:09:31 Voltage Offset Correction: v_off_ini = 2338.789 uV, v_off_fin = 2361.021 uV 14:09:31 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 126.870 uVs 14:09:31 Integrated Voltage Measurement... 14:09:40 Voltage Offset Correction: v_off_ini = 2368.854 uV, v_off_fin = 2361.076 uV 14:09:40 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -75.986 uVs 14:09:40 Integrated Voltage Measurement... 14:09:49 Voltage Offset Correction: v_off_ini = 2390.399 uV, v_off_fin = 2394.597 uV 14:09:49 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 45.718 uVs 14:09:53 I1X Average and RMS of Measurements VT = -80.636 +- 14.114 uVs I1X = 16.128 +- 2.823 uTm 14:09:56 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 14:10:00 Integrated Voltage Measurement... 14:10:09 Voltage Offset Correction: v_off_ini = 2376.007 uV, v_off_fin = 2384.703 uV 14:10:09 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -225.063 uVs 14:10:09 Integrated Voltage Measurement... 14:10:18 Voltage Offset Correction: v_off_ini = 2399.310 uV, v_off_fin = 2391.530 uV 14:10:18 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 246.377 uVs 14:10:18 Integrated Voltage Measurement... 14:10:27 Voltage Offset Correction: v_off_ini = 2391.877 uV, v_off_fin = 2385.349 uV 14:10:27 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -208.009 uVs 14:10:27 Integrated Voltage Measurement... 14:10:36 Voltage Offset Correction: v_off_ini = 2401.710 uV, v_off_fin = 2408.619 uV 14:10:36 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 247.727 uVs 14:10:36 Integrated Voltage Measurement... 14:10:45 Voltage Offset Correction: v_off_ini = 2389.701 uV, v_off_fin = 2372.505 uV 14:10:45 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -245.622 uVs 14:10:45 Integrated Voltage Measurement... 14:10:54 Voltage Offset Correction: v_off_ini = 2363.187 uV, v_off_fin = 2383.557 uV 14:10:54 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 244.003 uVs 14:10:58 I1Y Average and RMS of Measurements VT = -236.133 +- 6.924 uVs I1Y = -47.224 +- 1.387 uTm 14:11:03 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 14:11:06 Integrated Voltage Measurement... 14:11:15 Voltage Offset Correction: v_off_ini = 2374.506 uV, v_off_fin = 2370.695 uV 14:11:15 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -70.422 uVs 14:11:15 Integrated Voltage Measurement... 14:11:24 Voltage Offset Correction: v_off_ini = 2411.355 uV, v_off_fin = 2403.281 uV 14:11:24 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 68.268 uVs 14:11:24 Integrated Voltage Measurement... 14:11:33 Voltage Offset Correction: v_off_ini = 2360.481 uV, v_off_fin = 2394.203 uV 14:11:33 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -33.508 uVs 14:11:33 Integrated Voltage Measurement... 14:11:42 Voltage Offset Correction: v_off_ini = 2384.865 uV, v_off_fin = 2397.576 uV 14:11:42 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 90.028 uVs 14:11:42 Integrated Voltage Measurement... 14:11:51 Voltage Offset Correction: v_off_ini = 2375.891 uV, v_off_fin = 2353.310 uV 14:11:51 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -69.626 uVs 14:11:51 Integrated Voltage Measurement... 14:12:00 Voltage Offset Correction: v_off_ini = 2358.713 uV, v_off_fin = 2372.295 uV 14:12:00 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 102.060 uVs 14:12:02 I2X Average and RMS of Measurements VT = -72.319 +- 10.051 uVs I2X = 6.654 +- 0.925 uTm^2 14:12:07 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 14:12:10 Integrated Voltage Measurement... 14:12:19 Voltage Offset Correction: v_off_ini = 2432.258 uV, v_off_fin = 2392.425 uV 14:12:19 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -355.850 uVs 14:12:19 Integrated Voltage Measurement... 14:12:28 Voltage Offset Correction: v_off_ini = 2418.662 uV, v_off_fin = 2425.830 uV 14:12:28 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 358.674 uVs 14:12:28 Integrated Voltage Measurement... 14:12:37 Voltage Offset Correction: v_off_ini = 2423.018 uV, v_off_fin = 2425.199 uV 14:12:37 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -375.371 uVs 14:12:37 Integrated Voltage Measurement... 14:12:46 Voltage Offset Correction: v_off_ini = 2437.225 uV, v_off_fin = 2431.129 uV 14:12:46 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 336.129 uVs 14:12:46 Integrated Voltage Measurement... 14:12:55 Voltage Offset Correction: v_off_ini = 2429.087 uV, v_off_fin = 2443.310 uV 14:12:55 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -387.207 uVs 14:12:55 Integrated Voltage Measurement... 14:13:04 Voltage Offset Correction: v_off_ini = 2424.788 uV, v_off_fin = 2429.664 uV 14:13:04 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 372.619 uVs 14:13:06 I2Y Average and RMS of Measurements VT = -364.308 +- 11.051 uVs I2Y = -33.516 +- 1.017 uTm^2 14:13:17 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 14:13:21 Integrated Voltage Measurement... 14:13:30 Voltage Offset Correction: v_off_ini = 2437.836 uV, v_off_fin = 2414.957 uV 14:13:30 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -57.483 uVs 14:13:30 Integrated Voltage Measurement... 14:13:39 Voltage Offset Correction: v_off_ini = 2442.687 uV, v_off_fin = 2441.611 uV 14:13:39 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 58.003 uVs 14:13:40 Integrated Voltage Measurement... 14:13:48 Voltage Offset Correction: v_off_ini = 2440.298 uV, v_off_fin = 2440.985 uV 14:13:48 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -75.980 uVs 14:13:49 Integrated Voltage Measurement... 14:13:57 Voltage Offset Correction: v_off_ini = 2430.339 uV, v_off_fin = 2449.541 uV 14:13:57 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 90.976 uVs 14:13:58 Integrated Voltage Measurement... 14:14:06 Voltage Offset Correction: v_off_ini = 2441.413 uV, v_off_fin = 2444.564 uV 14:14:07 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -75.623 uVs 14:14:07 Integrated Voltage Measurement... 14:14:15 Voltage Offset Correction: v_off_ini = 2424.708 uV, v_off_fin = 2448.841 uV 14:14:16 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 78.239 uVs 14:14:20 I1X Average and RMS of Measurements VT = -72.717 +- 10.921 uVs I1X = 14.543 +- 2.184 uTm 14:14:22 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 14:14:26 Integrated Voltage Measurement... 14:14:35 Voltage Offset Correction: v_off_ini = 2455.534 uV, v_off_fin = 2435.078 uV 14:14:35 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -221.315 uVs 14:14:35 Integrated Voltage Measurement... 14:14:44 Voltage Offset Correction: v_off_ini = 2421.359 uV, v_off_fin = 2433.467 uV 14:14:44 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 243.080 uVs 14:14:44 Integrated Voltage Measurement... 14:14:53 Voltage Offset Correction: v_off_ini = 2418.256 uV, v_off_fin = 2421.537 uV 14:14:53 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -230.806 uVs 14:14:53 Integrated Voltage Measurement... 14:15:02 Voltage Offset Correction: v_off_ini = 2438.454 uV, v_off_fin = 2458.894 uV 14:15:02 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 232.264 uVs 14:15:02 Integrated Voltage Measurement... 14:15:11 Voltage Offset Correction: v_off_ini = 2435.690 uV, v_off_fin = 2406.347 uV 14:15:11 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -193.494 uVs 14:15:11 Integrated Voltage Measurement... 14:15:20 Voltage Offset Correction: v_off_ini = 2419.324 uV, v_off_fin = 2423.217 uV 14:15:20 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 220.048 uVs 14:15:25 I1Y Average and RMS of Measurements VT = -223.501 +- 11.833 uVs I1Y = -44.697 +- 2.367 uTm 14:15:30 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 14:15:32 Integrated Voltage Measurement... 14:15:41 Voltage Offset Correction: v_off_ini = 2451.440 uV, v_off_fin = 2427.758 uV 14:15:41 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -145.257 uVs 14:15:41 Integrated Voltage Measurement... 14:15:50 Voltage Offset Correction: v_off_ini = 2457.864 uV, v_off_fin = 2486.147 uV 14:15:50 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 47.547 uVs 14:15:50 Integrated Voltage Measurement... 14:15:59 Voltage Offset Correction: v_off_ini = 2453.189 uV, v_off_fin = 2428.551 uV 14:15:59 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -59.691 uVs 14:15:59 Integrated Voltage Measurement... 14:16:08 Voltage Offset Correction: v_off_ini = 2431.892 uV, v_off_fin = 2431.978 uV 14:16:08 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 68.005 uVs 14:16:08 Integrated Voltage Measurement... 14:16:17 Voltage Offset Correction: v_off_ini = 2416.367 uV, v_off_fin = 2423.475 uV 14:16:17 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -78.724 uVs 14:16:17 Integrated Voltage Measurement... 14:16:26 Voltage Offset Correction: v_off_ini = 2424.522 uV, v_off_fin = 2434.686 uV 14:16:26 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 71.724 uVs 14:16:29 I2X Average and RMS of Measurements VT = -78.491 +- 13.489 uVs I2X = 7.221 +- 1.241 uTm^2 14:16:34 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 14:16:36 Integrated Voltage Measurement... 14:16:45 Voltage Offset Correction: v_off_ini = 2392.856 uV, v_off_fin = 2392.004 uV 14:16:45 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -357.607 uVs 14:16:45 Integrated Voltage Measurement... 14:16:54 Voltage Offset Correction: v_off_ini = 2380.064 uV, v_off_fin = 2387.832 uV 14:16:54 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 340.949 uVs 14:16:54 Integrated Voltage Measurement... 14:17:03 Voltage Offset Correction: v_off_ini = 2381.512 uV, v_off_fin = 2363.701 uV 14:17:03 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -358.995 uVs 14:17:03 Integrated Voltage Measurement... 14:17:12 Voltage Offset Correction: v_off_ini = 2395.426 uV, v_off_fin = 2398.607 uV 14:17:12 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 372.501 uVs 14:17:13 Integrated Voltage Measurement... 14:17:21 Voltage Offset Correction: v_off_ini = 2412.247 uV, v_off_fin = 2389.238 uV 14:17:21 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -361.005 uVs 14:17:22 Integrated Voltage Measurement... 14:17:31 Voltage Offset Correction: v_off_ini = 2395.951 uV, v_off_fin = 2381.058 uV 14:17:31 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 375.305 uVs 14:17:33 I2Y Average and RMS of Measurements VT = -361.060 +- 8.389 uVs I2Y = -33.218 +- 0.772 uTm^2 14:17:44 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 14:17:48 Integrated Voltage Measurement... 14:17:57 Voltage Offset Correction: v_off_ini = 2400.969 uV, v_off_fin = 2374.807 uV 14:17:57 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -66.621 uVs 14:17:57 Integrated Voltage Measurement... 14:18:06 Voltage Offset Correction: v_off_ini = 2370.139 uV, v_off_fin = 2367.415 uV 14:18:06 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 79.030 uVs 14:18:06 Integrated Voltage Measurement... 14:18:15 Voltage Offset Correction: v_off_ini = 2376.603 uV, v_off_fin = 2363.513 uV 14:18:15 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -50.552 uVs 14:18:15 Integrated Voltage Measurement... 14:18:24 Voltage Offset Correction: v_off_ini = 2377.762 uV, v_off_fin = 2374.396 uV 14:18:24 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 73.303 uVs 14:18:24 Integrated Voltage Measurement... 14:18:33 Voltage Offset Correction: v_off_ini = 2360.989 uV, v_off_fin = 2370.504 uV 14:18:33 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -73.600 uVs 14:18:33 Integrated Voltage Measurement... 14:18:42 Voltage Offset Correction: v_off_ini = 2345.360 uV, v_off_fin = 2379.018 uV 14:18:42 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 97.296 uVs 14:18:46 I1X Average and RMS of Measurements VT = -73.400 +- 9.611 uVs I1X = 14.680 +- 1.922 uTm 14:18:49 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 14:18:53 Integrated Voltage Measurement... 14:19:02 Voltage Offset Correction: v_off_ini = 2345.385 uV, v_off_fin = 2341.463 uV 14:19:02 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -213.408 uVs 14:19:02 Integrated Voltage Measurement... 14:19:11 Voltage Offset Correction: v_off_ini = 2371.795 uV, v_off_fin = 2380.579 uV 14:19:11 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 209.576 uVs 14:19:11 Integrated Voltage Measurement... 14:19:20 Voltage Offset Correction: v_off_ini = 2332.708 uV, v_off_fin = 2350.804 uV 14:19:20 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -210.179 uVs 14:19:20 Integrated Voltage Measurement... 14:19:29 Voltage Offset Correction: v_off_ini = 2382.165 uV, v_off_fin = 2368.213 uV 14:19:29 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 202.298 uVs 14:19:29 Integrated Voltage Measurement... 14:19:38 Voltage Offset Correction: v_off_ini = 2362.363 uV, v_off_fin = 2356.127 uV 14:19:38 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -227.170 uVs 14:19:38 Integrated Voltage Measurement... 14:19:47 Voltage Offset Correction: v_off_ini = 2342.468 uV, v_off_fin = 2380.226 uV 14:19:47 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 224.435 uVs 14:19:51 I1Y Average and RMS of Measurements VT = -214.511 +- 8.267 uVs I1Y = -42.899 +- 1.652 uTm