SLAC Magnetic Measurements Date: 07-15-2026 Time: 15:58:18 Project: LCLS-II-HE Magnet Type: Phase Shifter Magnet Name: HE_SXPS_16335 Dataset Number: 1 Run Purpose: Tuning Run Type: Run Number: 3 Measurement Device: Stretched Wire Operator: YL Comment: 15:59:16 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 15:59:19 Motion Parameters, X Controller, Axis 1 acc = 0.001000 m/s/s dcl = 0.001000 m/s/s vel = 0.001000 m/s pos = 0.000000 m kp = 188.600000 ki = 820.000000 kd = 491.400000 db = 0.000000 15:59:20 Motion Parameters, X Controller, Axis 2 acc = 0.001000 m/s/s dcl = 0.001000 m/s/s vel = 0.001000 m/s pos = 0.000000 m kp = 188.600000 ki = 820.000000 kd = 491.400000 db = 0.000000 15:59:22 Motion Parameters, Y Controller, Axis 1 acc = 0.001000 m/s/s dcl = 0.001000 m/s/s vel = 0.001000 m/s pos = 0.000000 m kp = 188.600000 ki = 820.000000 kd = 491.400000 db = 0.000000 15:59:23 Motion Parameters, Y Controller, Axis 2 acc = 0.001000 m/s/s dcl = 0.001000 m/s/s vel = 0.001000 m/s pos = 0.000000 m kp = 188.600000 ki = 820.000000 kd = 491.400000 db = 0.000000 15:59:38 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 15:59:42 Integrated Voltage Measurement... 15:59:51 Voltage Offset Correction: v_off_ini = 2398.788 uV, v_off_fin = 2383.644 uV 15:59:51 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.4998 mm dXY = -0.4999 mm VT = -121.597 uVs 15:59:52 Integrated Voltage Measurement... 16:00:00 Voltage Offset Correction: v_off_ini = 2397.703 uV, v_off_fin = 2398.762 uV 16:00:00 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = 99.577 uVs 16:00:01 Integrated Voltage Measurement... 16:00:09 Voltage Offset Correction: v_off_ini = 2385.223 uV, v_off_fin = 2384.965 uV 16:00:09 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -100.785 uVs 16:00:10 Integrated Voltage Measurement... 16:00:18 Voltage Offset Correction: v_off_ini = 2388.471 uV, v_off_fin = 2377.939 uV 16:00:18 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 88.352 uVs 16:00:19 Integrated Voltage Measurement... 16:00:27 Voltage Offset Correction: v_off_ini = 2401.841 uV, v_off_fin = 2408.588 uV 16:00:27 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -97.942 uVs 16:00:28 Integrated Voltage Measurement... 16:00:36 Voltage Offset Correction: v_off_ini = 2405.432 uV, v_off_fin = 2364.699 uV 16:00:36 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 84.733 uVs 16:00:41 I1X Average and RMS of Measurements VT = -98.831 +- 8.417 uVs I1X = 19.768 +- 1.685 uTm 16:00:43 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:00:47 Integrated Voltage Measurement... 16:00:56 Voltage Offset Correction: v_off_ini = 2391.376 uV, v_off_fin = 2363.769 uV 16:00:56 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -270.380 uVs 16:00:56 Integrated Voltage Measurement... 16:01:05 Voltage Offset Correction: v_off_ini = 2385.929 uV, v_off_fin = 2399.172 uV 16:01:05 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 288.751 uVs 16:01:05 Integrated Voltage Measurement... 16:01:14 Voltage Offset Correction: v_off_ini = 2372.813 uV, v_off_fin = 2380.011 uV 16:01:14 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -268.758 uVs 16:01:14 Integrated Voltage Measurement... 16:01:23 Voltage Offset Correction: v_off_ini = 2401.275 uV, v_off_fin = 2412.361 uV 16:01:23 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 282.834 uVs 16:01:23 Integrated Voltage Measurement... 16:01:32 Voltage Offset Correction: v_off_ini = 2388.803 uV, v_off_fin = 2376.477 uV 16:01:32 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -278.351 uVs 16:01:32 Integrated Voltage Measurement... 16:01:41 Voltage Offset Correction: v_off_ini = 2415.888 uV, v_off_fin = 2439.253 uV 16:01:41 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 279.139 uVs 16:01:46 I1Y Average and RMS of Measurements VT = -278.036 +- 1.619 uVs I1Y = -55.607 +- 0.324 uTm 16:01:51 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:01:53 Integrated Voltage Measurement... 16:02:02 Voltage Offset Correction: v_off_ini = 2415.088 uV, v_off_fin = 2409.028 uV 16:02:02 Integrated Voltage Measurement dXYus = -1.0002 mm, dXYds = 0.0000 mm dXY = -1.0002 mm VT = -29.391 uVs 16:02:02 Integrated Voltage Measurement... 16:02:11 Voltage Offset Correction: v_off_ini = 2420.945 uV, v_off_fin = 2430.184 uV 16:02:11 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 20.513 uVs 16:02:11 Integrated Voltage Measurement... 16:02:20 Voltage Offset Correction: v_off_ini = 2395.365 uV, v_off_fin = 2406.129 uV 16:02:20 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -26.619 uVs 16:02:20 Integrated Voltage Measurement... 16:02:29 Voltage Offset Correction: v_off_ini = 2415.943 uV, v_off_fin = 2432.359 uV 16:02:29 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 39.599 uVs 16:02:29 Integrated Voltage Measurement... 16:02:38 Voltage Offset Correction: v_off_ini = 2409.137 uV, v_off_fin = 2410.649 uV 16:02:38 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -51.304 uVs 16:02:38 Integrated Voltage Measurement... 16:02:47 Voltage Offset Correction: v_off_ini = 2422.731 uV, v_off_fin = 2420.717 uV 16:02:47 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 12.592 uVs 16:02:50 I2X Average and RMS of Measurements VT = -30.003 +- 3.603 uVs I2X = 2.760 +- 0.332 uTm^2 16:02:55 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:02:57 Integrated Voltage Measurement... 16:03:06 Voltage Offset Correction: v_off_ini = 2417.799 uV, v_off_fin = 2408.942 uV 16:03:06 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -288.933 uVs 16:03:06 Integrated Voltage Measurement... 16:03:15 Voltage Offset Correction: v_off_ini = 2418.901 uV, v_off_fin = 2425.433 uV 16:03:15 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 284.798 uVs 16:03:15 Integrated Voltage Measurement... 16:03:24 Voltage Offset Correction: v_off_ini = 2427.302 uV, v_off_fin = 2421.355 uV 16:03:24 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -265.577 uVs 16:03:24 Integrated Voltage Measurement... 16:03:33 Voltage Offset Correction: v_off_ini = 2431.007 uV, v_off_fin = 2425.072 uV 16:03:33 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 282.121 uVs 16:03:33 Integrated Voltage Measurement... 16:03:42 Voltage Offset Correction: v_off_ini = 2417.869 uV, v_off_fin = 2439.387 uV 16:03:42 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -278.647 uVs 16:03:42 Integrated Voltage Measurement... 16:03:51 Voltage Offset Correction: v_off_ini = 2404.921 uV, v_off_fin = 2412.670 uV 16:03:51 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 291.203 uVs 16:03:54 I2Y Average and RMS of Measurements VT = -281.880 +- 5.734 uVs I2Y = -25.933 +- 0.528 uTm^2 16:04:04 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:04:09 Integrated Voltage Measurement... 16:04:18 Voltage Offset Correction: v_off_ini = 2391.730 uV, v_off_fin = 2394.357 uV 16:04:18 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = -108.378 uVs 16:04:18 Integrated Voltage Measurement... 16:04:27 Voltage Offset Correction: v_off_ini = 2391.323 uV, v_off_fin = 2399.898 uV 16:04:27 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = 103.134 uVs 16:04:27 Integrated Voltage Measurement... 16:04:36 Voltage Offset Correction: v_off_ini = 2407.723 uV, v_off_fin = 2407.993 uV 16:04:36 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -92.938 uVs 16:04:36 Integrated Voltage Measurement... 16:04:45 Voltage Offset Correction: v_off_ini = 2384.282 uV, v_off_fin = 2385.843 uV 16:04:45 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 127.139 uVs 16:04:45 Integrated Voltage Measurement... 16:04:54 Voltage Offset Correction: v_off_ini = 2384.920 uV, v_off_fin = 2405.394 uV 16:04:54 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -116.119 uVs 16:04:54 Integrated Voltage Measurement... 16:05:03 Voltage Offset Correction: v_off_ini = 2388.483 uV, v_off_fin = 2380.052 uV 16:05:03 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 105.519 uVs 16:05:07 I1X Average and RMS of Measurements VT = -108.871 +- 2.226 uVs I1X = 21.774 +- 0.445 uTm 16:05:09 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:05:14 Integrated Voltage Measurement... 16:05:23 Voltage Offset Correction: v_off_ini = 2354.192 uV, v_off_fin = 2355.083 uV 16:05:23 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -272.890 uVs 16:05:23 Integrated Voltage Measurement... 16:05:31 Voltage Offset Correction: v_off_ini = 2370.248 uV, v_off_fin = 2399.440 uV 16:05:32 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 269.106 uVs 16:05:32 Integrated Voltage Measurement... 16:05:40 Voltage Offset Correction: v_off_ini = 2374.978 uV, v_off_fin = 2344.909 uV 16:05:40 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -277.992 uVs 16:05:40 Integrated Voltage Measurement... 16:05:49 Voltage Offset Correction: v_off_ini = 2369.746 uV, v_off_fin = 2359.274 uV 16:05:49 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 253.419 uVs 16:05:49 Integrated Voltage Measurement... 16:05:58 Voltage Offset Correction: v_off_ini = 2358.721 uV, v_off_fin = 2357.418 uV 16:05:58 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -230.479 uVs 16:05:58 Integrated Voltage Measurement... 16:06:07 Voltage Offset Correction: v_off_ini = 2357.590 uV, v_off_fin = 2363.292 uV 16:06:07 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 234.509 uVs 16:06:12 I1Y Average and RMS of Measurements VT = -256.399 +- 17.041 uVs I1Y = -51.280 +- 3.408 uTm 16:06:17 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:06:19 Integrated Voltage Measurement... 16:06:28 Voltage Offset Correction: v_off_ini = 2390.312 uV, v_off_fin = 2394.668 uV 16:06:28 Integrated Voltage Measurement dXYus = -1.0002 mm, dXYds = 0.0000 mm dXY = -1.0002 mm VT = -51.305 uVs 16:06:28 Integrated Voltage Measurement... 16:06:37 Voltage Offset Correction: v_off_ini = 2392.728 uV, v_off_fin = 2416.390 uV 16:06:37 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 59.981 uVs 16:06:37 Integrated Voltage Measurement... 16:06:46 Voltage Offset Correction: v_off_ini = 2431.216 uV, v_off_fin = 2429.279 uV 16:06:46 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -39.131 uVs 16:06:46 Integrated Voltage Measurement... 16:06:55 Voltage Offset Correction: v_off_ini = 2439.910 uV, v_off_fin = 2436.714 uV 16:06:55 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 55.009 uVs 16:06:55 Integrated Voltage Measurement... 16:07:04 Voltage Offset Correction: v_off_ini = 2406.765 uV, v_off_fin = 2400.074 uV 16:07:04 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -28.750 uVs 16:07:04 Integrated Voltage Measurement... 16:07:13 Voltage Offset Correction: v_off_ini = 2415.262 uV, v_off_fin = 2437.898 uV 16:07:13 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 61.959 uVs 16:07:16 I2X Average and RMS of Measurements VT = -49.356 +- 4.500 uVs I2X = 4.541 +- 0.414 uTm^2 16:07:21 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:07:23 Integrated Voltage Measurement... 16:07:32 Voltage Offset Correction: v_off_ini = 2431.631 uV, v_off_fin = 2422.257 uV 16:07:32 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -277.018 uVs 16:07:32 Integrated Voltage Measurement... 16:07:41 Voltage Offset Correction: v_off_ini = 2428.539 uV, v_off_fin = 2459.513 uV 16:07:41 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 273.708 uVs 16:07:41 Integrated Voltage Measurement... 16:07:50 Voltage Offset Correction: v_off_ini = 2454.856 uV, v_off_fin = 2440.124 uV 16:07:50 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -286.630 uVs 16:07:50 Integrated Voltage Measurement... 16:07:59 Voltage Offset Correction: v_off_ini = 2416.197 uV, v_off_fin = 2445.200 uV 16:07:59 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 277.349 uVs 16:07:59 Integrated Voltage Measurement... 16:08:08 Voltage Offset Correction: v_off_ini = 2437.312 uV, v_off_fin = 2439.893 uV 16:08:08 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = -269.612 uVs 16:08:08 Integrated Voltage Measurement... 16:08:17 Voltage Offset Correction: v_off_ini = 2456.326 uV, v_off_fin = 2451.263 uV 16:08:17 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 279.629 uVs 16:08:20 I2Y Average and RMS of Measurements VT = -277.324 +- 3.313 uVs I2Y = -25.514 +- 0.304 uTm^2 16:08:30 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:08:35 Integrated Voltage Measurement... 16:08:44 Voltage Offset Correction: v_off_ini = 2455.990 uV, v_off_fin = 2441.458 uV 16:08:44 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.4999 mm dXY = -0.4999 mm VT = -119.855 uVs 16:08:44 Integrated Voltage Measurement... 16:08:53 Voltage Offset Correction: v_off_ini = 2459.159 uV, v_off_fin = 2464.016 uV 16:08:53 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 145.712 uVs 16:08:53 Integrated Voltage Measurement... 16:09:02 Voltage Offset Correction: v_off_ini = 2472.111 uV, v_off_fin = 2479.519 uV 16:09:02 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -97.861 uVs 16:09:02 Integrated Voltage Measurement... 16:09:11 Voltage Offset Correction: v_off_ini = 2477.945 uV, v_off_fin = 2490.343 uV 16:09:11 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = 135.223 uVs 16:09:11 Integrated Voltage Measurement... 16:09:20 Voltage Offset Correction: v_off_ini = 2488.609 uV, v_off_fin = 2482.913 uV 16:09:20 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -97.839 uVs 16:09:20 Integrated Voltage Measurement... 16:09:29 Voltage Offset Correction: v_off_ini = 2509.359 uV, v_off_fin = 2477.830 uV 16:09:29 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 91.899 uVs 16:09:33 I1X Average and RMS of Measurements VT = -114.732 +- 15.531 uVs I1X = 22.948 +- 3.107 uTm 16:09:35 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:09:40 Integrated Voltage Measurement... 16:09:48 Voltage Offset Correction: v_off_ini = 2485.583 uV, v_off_fin = 2475.672 uV 16:09:49 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -241.394 uVs 16:09:49 Integrated Voltage Measurement... 16:09:57 Voltage Offset Correction: v_off_ini = 2477.060 uV, v_off_fin = 2485.082 uV 16:09:57 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 247.810 uVs 16:09:58 Integrated Voltage Measurement... 16:10:06 Voltage Offset Correction: v_off_ini = 2475.392 uV, v_off_fin = 2486.370 uV 16:10:06 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -226.672 uVs 16:10:06 Integrated Voltage Measurement... 16:10:15 Voltage Offset Correction: v_off_ini = 2474.802 uV, v_off_fin = 2485.823 uV 16:10:15 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 262.998 uVs 16:10:15 Integrated Voltage Measurement... 16:10:24 Voltage Offset Correction: v_off_ini = 2471.345 uV, v_off_fin = 2467.933 uV 16:10:24 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -230.750 uVs 16:10:24 Integrated Voltage Measurement... 16:10:33 Voltage Offset Correction: v_off_ini = 2515.215 uV, v_off_fin = 2509.037 uV 16:10:33 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 263.122 uVs 16:10:38 I1Y Average and RMS of Measurements VT = -245.457 +- 1.050 uVs I1Y = -49.092 +- 0.209 uTm 16:10:43 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:10:45 Integrated Voltage Measurement... 16:10:54 Voltage Offset Correction: v_off_ini = 2488.567 uV, v_off_fin = 2481.385 uV 16:10:54 Integrated Voltage Measurement dXYus = -1.0002 mm, dXYds = 0.0000 mm dXY = -1.0002 mm VT = -79.101 uVs 16:10:54 Integrated Voltage Measurement... 16:11:03 Voltage Offset Correction: v_off_ini = 2499.652 uV, v_off_fin = 2496.172 uV 16:11:03 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 83.511 uVs 16:11:03 Integrated Voltage Measurement... 16:11:12 Voltage Offset Correction: v_off_ini = 2518.626 uV, v_off_fin = 2495.746 uV 16:11:12 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -96.464 uVs 16:11:12 Integrated Voltage Measurement... 16:11:21 Voltage Offset Correction: v_off_ini = 2530.746 uV, v_off_fin = 2539.782 uV 16:11:21 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 74.388 uVs 16:11:21 Integrated Voltage Measurement... 16:11:30 Voltage Offset Correction: v_off_ini = 2531.455 uV, v_off_fin = 2515.336 uV 16:11:30 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -53.905 uVs 16:11:30 Integrated Voltage Measurement... 16:11:39 Voltage Offset Correction: v_off_ini = 2520.817 uV, v_off_fin = 2530.688 uV 16:11:39 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 67.197 uVs 16:11:42 I2X Average and RMS of Measurements VT = -75.761 +- 10.886 uVs I2X = 6.970 +- 1.001 uTm^2 16:11:47 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:11:49 Integrated Voltage Measurement... 16:11:58 Voltage Offset Correction: v_off_ini = 2504.567 uV, v_off_fin = 2498.892 uV 16:11:58 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -255.340 uVs 16:11:58 Integrated Voltage Measurement... 16:12:07 Voltage Offset Correction: v_off_ini = 2490.567 uV, v_off_fin = 2506.654 uV 16:12:07 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 268.921 uVs 16:12:07 Integrated Voltage Measurement... 16:12:16 Voltage Offset Correction: v_off_ini = 2479.248 uV, v_off_fin = 2498.652 uV 16:12:16 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -256.100 uVs 16:12:16 Integrated Voltage Measurement... 16:12:25 Voltage Offset Correction: v_off_ini = 2509.434 uV, v_off_fin = 2505.150 uV 16:12:25 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 253.715 uVs 16:12:25 Integrated Voltage Measurement... 16:12:34 Voltage Offset Correction: v_off_ini = 2507.890 uV, v_off_fin = 2508.519 uV 16:12:34 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -248.847 uVs 16:12:34 Integrated Voltage Measurement... 16:12:43 Voltage Offset Correction: v_off_ini = 2508.130 uV, v_off_fin = 2489.135 uV 16:12:43 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 242.335 uVs 16:12:46 I2Y Average and RMS of Measurements VT = -254.210 +- 6.770 uVs I2Y = -23.387 +- 0.623 uTm^2 16:12:57 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:13:01 Integrated Voltage Measurement... 16:13:10 Voltage Offset Correction: v_off_ini = 2501.477 uV, v_off_fin = 2498.218 uV 16:13:10 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.4998 mm dXY = -0.4999 mm VT = -123.438 uVs 16:13:10 Integrated Voltage Measurement... 16:13:19 Voltage Offset Correction: v_off_ini = 2534.488 uV, v_off_fin = 2519.828 uV 16:13:19 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 113.694 uVs 16:13:19 Integrated Voltage Measurement... 16:13:28 Voltage Offset Correction: v_off_ini = 2549.664 uV, v_off_fin = 2528.934 uV 16:13:28 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -145.006 uVs 16:13:28 Integrated Voltage Measurement... 16:13:37 Voltage Offset Correction: v_off_ini = 2533.407 uV, v_off_fin = 2501.555 uV 16:13:37 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 141.160 uVs 16:13:37 Integrated Voltage Measurement... 16:13:46 Voltage Offset Correction: v_off_ini = 2497.724 uV, v_off_fin = 2521.507 uV 16:13:46 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -132.320 uVs 16:13:46 Integrated Voltage Measurement... 16:13:55 Voltage Offset Correction: v_off_ini = 2529.287 uV, v_off_fin = 2542.166 uV 16:13:55 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 153.513 uVs 16:13:59 I1X Average and RMS of Measurements VT = -134.855 +- 11.518 uVs I1X = 26.974 +- 2.303 uTm 16:14:01 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:14:06 Integrated Voltage Measurement... 16:14:15 Voltage Offset Correction: v_off_ini = 2532.003 uV, v_off_fin = 2525.337 uV 16:14:15 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -223.340 uVs 16:14:15 Integrated Voltage Measurement... 16:14:24 Voltage Offset Correction: v_off_ini = 2541.313 uV, v_off_fin = 2558.155 uV 16:14:24 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 228.223 uVs 16:14:24 Integrated Voltage Measurement... 16:14:33 Voltage Offset Correction: v_off_ini = 2538.543 uV, v_off_fin = 2534.816 uV 16:14:33 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -232.286 uVs 16:14:33 Integrated Voltage Measurement... 16:14:42 Voltage Offset Correction: v_off_ini = 2540.036 uV, v_off_fin = 2568.618 uV 16:14:42 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 239.930 uVs 16:14:42 Integrated Voltage Measurement... 16:14:51 Voltage Offset Correction: v_off_ini = 2560.082 uV, v_off_fin = 2544.585 uV 16:14:51 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -224.336 uVs 16:14:51 Integrated Voltage Measurement... 16:15:00 Voltage Offset Correction: v_off_ini = 2543.098 uV, v_off_fin = 2571.561 uV 16:15:00 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 234.822 uVs 16:15:04 I1Y Average and RMS of Measurements VT = -230.490 +- 4.265 uVs I1Y = -46.098 +- 0.853 uTm 16:15:09 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:15:11 Integrated Voltage Measurement... 16:15:20 Voltage Offset Correction: v_off_ini = 2602.860 uV, v_off_fin = 2572.082 uV 16:15:20 Integrated Voltage Measurement dXYus = -1.0001 mm, dXYds = 0.0000 mm dXY = -1.0001 mm VT = -110.390 uVs 16:15:20 Integrated Voltage Measurement... 16:15:29 Voltage Offset Correction: v_off_ini = 2552.665 uV, v_off_fin = 2563.492 uV 16:15:29 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 66.629 uVs 16:15:29 Integrated Voltage Measurement... 16:15:38 Voltage Offset Correction: v_off_ini = 2545.343 uV, v_off_fin = 2561.958 uV 16:15:38 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -46.283 uVs 16:15:39 Integrated Voltage Measurement... 16:15:47 Voltage Offset Correction: v_off_ini = 2597.887 uV, v_off_fin = 2570.528 uV 16:15:48 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 63.233 uVs 16:15:48 Integrated Voltage Measurement... 16:15:57 Voltage Offset Correction: v_off_ini = 2553.230 uV, v_off_fin = 2529.896 uV 16:15:57 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -85.232 uVs 16:15:57 Integrated Voltage Measurement... 16:16:06 Voltage Offset Correction: v_off_ini = 2542.091 uV, v_off_fin = 2550.629 uV 16:16:06 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 55.840 uVs 16:16:08 I2X Average and RMS of Measurements VT = -71.268 +- 13.789 uVs I2X = 6.557 +- 1.268 uTm^2 16:16:13 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:16:16 Integrated Voltage Measurement... 16:16:25 Voltage Offset Correction: v_off_ini = 2569.797 uV, v_off_fin = 2537.983 uV 16:16:25 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = -237.856 uVs 16:16:25 Integrated Voltage Measurement... 16:16:34 Voltage Offset Correction: v_off_ini = 2542.639 uV, v_off_fin = 2555.043 uV 16:16:34 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 260.676 uVs 16:16:34 Integrated Voltage Measurement... 16:16:43 Voltage Offset Correction: v_off_ini = 2577.220 uV, v_off_fin = 2588.949 uV 16:16:43 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -228.011 uVs 16:16:43 Integrated Voltage Measurement... 16:16:52 Voltage Offset Correction: v_off_ini = 2577.674 uV, v_off_fin = 2565.085 uV 16:16:52 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 232.390 uVs 16:16:52 Integrated Voltage Measurement... 16:17:01 Voltage Offset Correction: v_off_ini = 2559.190 uV, v_off_fin = 2558.551 uV 16:17:01 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -220.648 uVs 16:17:01 Integrated Voltage Measurement... 16:17:10 Voltage Offset Correction: v_off_ini = 2541.260 uV, v_off_fin = 2541.447 uV 16:17:10 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 245.034 uVs 16:17:12 I2Y Average and RMS of Measurements VT = -237.436 +- 8.434 uVs I2Y = -21.844 +- 0.776 uTm^2 16:17:23 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:17:27 Integrated Voltage Measurement... 16:17:36 Voltage Offset Correction: v_off_ini = 2546.466 uV, v_off_fin = 2530.403 uV 16:17:36 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = -167.460 uVs 16:17:36 Integrated Voltage Measurement... 16:17:45 Voltage Offset Correction: v_off_ini = 2559.683 uV, v_off_fin = 2556.261 uV 16:17:45 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 135.839 uVs 16:17:45 Integrated Voltage Measurement... 16:17:54 Voltage Offset Correction: v_off_ini = 2577.302 uV, v_off_fin = 2571.382 uV 16:17:54 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -95.472 uVs 16:17:54 Integrated Voltage Measurement... 16:18:03 Voltage Offset Correction: v_off_ini = 2526.722 uV, v_off_fin = 2545.258 uV 16:18:03 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 135.130 uVs 16:18:03 Integrated Voltage Measurement... 16:18:12 Voltage Offset Correction: v_off_ini = 2535.016 uV, v_off_fin = 2540.867 uV 16:18:12 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -143.680 uVs 16:18:12 Integrated Voltage Measurement... 16:18:21 Voltage Offset Correction: v_off_ini = 2561.151 uV, v_off_fin = 2578.660 uV 16:18:21 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 143.193 uVs 16:18:25 I1X Average and RMS of Measurements VT = -136.796 +- 15.564 uVs I1X = 27.359 +- 3.113 uTm 16:18:28 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:18:32 Integrated Voltage Measurement... 16:18:41 Voltage Offset Correction: v_off_ini = 2574.445 uV, v_off_fin = 2550.915 uV 16:18:41 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -186.672 uVs 16:18:41 Integrated Voltage Measurement... 16:18:50 Voltage Offset Correction: v_off_ini = 2566.713 uV, v_off_fin = 2596.812 uV 16:18:50 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 245.442 uVs 16:18:50 Integrated Voltage Measurement... 16:18:59 Voltage Offset Correction: v_off_ini = 2535.262 uV, v_off_fin = 2533.206 uV 16:18:59 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -212.439 uVs 16:18:59 Integrated Voltage Measurement... 16:19:08 Voltage Offset Correction: v_off_ini = 2571.091 uV, v_off_fin = 2566.607 uV 16:19:08 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 243.057 uVs 16:19:08 Integrated Voltage Measurement... 16:19:17 Voltage Offset Correction: v_off_ini = 2539.895 uV, v_off_fin = 2541.437 uV 16:19:17 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -212.232 uVs 16:19:17 Integrated Voltage Measurement... 16:19:26 Voltage Offset Correction: v_off_ini = 2591.390 uV, v_off_fin = 2571.824 uV 16:19:26 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 231.477 uVs 16:19:30 I1Y Average and RMS of Measurements VT = -221.886 +- 4.773 uVs I1Y = -44.377 +- 0.955 uTm 16:19:35 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:19:38 Integrated Voltage Measurement... 16:19:47 Voltage Offset Correction: v_off_ini = 2579.760 uV, v_off_fin = 2581.312 uV 16:19:47 Integrated Voltage Measurement dXYus = -1.0002 mm, dXYds = 0.0000 mm dXY = -1.0002 mm VT = -75.681 uVs 16:19:47 Integrated Voltage Measurement... 16:19:56 Voltage Offset Correction: v_off_ini = 2595.454 uV, v_off_fin = 2601.007 uV 16:19:56 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 88.216 uVs 16:19:56 Integrated Voltage Measurement... 16:20:05 Voltage Offset Correction: v_off_ini = 2573.107 uV, v_off_fin = 2563.954 uV 16:20:05 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -66.644 uVs 16:20:05 Integrated Voltage Measurement... 16:20:14 Voltage Offset Correction: v_off_ini = 2588.304 uV, v_off_fin = 2607.179 uV 16:20:14 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 79.930 uVs 16:20:14 Integrated Voltage Measurement... 16:20:23 Voltage Offset Correction: v_off_ini = 2582.434 uV, v_off_fin = 2580.345 uV 16:20:23 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -72.292 uVs 16:20:23 Integrated Voltage Measurement... 16:20:32 Voltage Offset Correction: v_off_ini = 2562.136 uV, v_off_fin = 2575.928 uV 16:20:32 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 56.625 uVs 16:20:35 I2X Average and RMS of Measurements VT = -73.231 +- 7.141 uVs I2X = 6.737 +- 0.657 uTm^2 16:20:40 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:20:42 Integrated Voltage Measurement... 16:20:51 Voltage Offset Correction: v_off_ini = 2601.389 uV, v_off_fin = 2585.138 uV 16:20:51 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -206.420 uVs 16:20:51 Integrated Voltage Measurement... 16:21:00 Voltage Offset Correction: v_off_ini = 2572.401 uV, v_off_fin = 2586.255 uV 16:21:00 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 189.387 uVs 16:21:00 Integrated Voltage Measurement... 16:21:09 Voltage Offset Correction: v_off_ini = 2612.343 uV, v_off_fin = 2595.420 uV 16:21:09 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -203.448 uVs 16:21:09 Integrated Voltage Measurement... 16:21:18 Voltage Offset Correction: v_off_ini = 2577.919 uV, v_off_fin = 2578.641 uV 16:21:18 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 234.443 uVs 16:21:18 Integrated Voltage Measurement... 16:21:27 Voltage Offset Correction: v_off_ini = 2589.467 uV, v_off_fin = 2626.002 uV 16:21:27 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -198.681 uVs 16:21:28 Integrated Voltage Measurement... 16:21:36 Voltage Offset Correction: v_off_ini = 2610.622 uV, v_off_fin = 2621.929 uV 16:21:37 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 225.346 uVs 16:21:39 I2Y Average and RMS of Measurements VT = -209.621 +- 8.755 uVs I2Y = -19.285 +- 0.806 uTm^2 16:21:50 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:21:54 Integrated Voltage Measurement... 16:22:03 Voltage Offset Correction: v_off_ini = 2650.048 uV, v_off_fin = 2652.468 uV 16:22:03 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = -129.797 uVs 16:22:03 Integrated Voltage Measurement... 16:22:12 Voltage Offset Correction: v_off_ini = 2631.535 uV, v_off_fin = 2637.776 uV 16:22:12 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 149.981 uVs 16:22:12 Integrated Voltage Measurement... 16:22:21 Voltage Offset Correction: v_off_ini = 2645.262 uV, v_off_fin = 2639.551 uV 16:22:21 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -135.370 uVs 16:22:21 Integrated Voltage Measurement... 16:22:30 Voltage Offset Correction: v_off_ini = 2625.256 uV, v_off_fin = 2643.456 uV 16:22:30 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 156.611 uVs 16:22:30 Integrated Voltage Measurement... 16:22:39 Voltage Offset Correction: v_off_ini = 2626.436 uV, v_off_fin = 2629.772 uV 16:22:39 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -147.948 uVs 16:22:39 Integrated Voltage Measurement... 16:22:48 Voltage Offset Correction: v_off_ini = 2633.706 uV, v_off_fin = 2627.413 uV 16:22:48 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 181.497 uVs 16:22:52 I1X Average and RMS of Measurements VT = -150.201 +- 10.566 uVs I1X = 30.041 +- 2.113 uTm 16:22:55 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:22:59 Integrated Voltage Measurement... 16:23:08 Voltage Offset Correction: v_off_ini = 2666.612 uV, v_off_fin = 2672.247 uV 16:23:08 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -212.061 uVs 16:23:08 Integrated Voltage Measurement... 16:23:17 Voltage Offset Correction: v_off_ini = 2674.349 uV, v_off_fin = 2674.915 uV 16:23:17 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 245.579 uVs 16:23:17 Integrated Voltage Measurement... 16:23:26 Voltage Offset Correction: v_off_ini = 2644.387 uV, v_off_fin = 2638.317 uV 16:23:26 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -236.614 uVs 16:23:26 Integrated Voltage Measurement... 16:23:35 Voltage Offset Correction: v_off_ini = 2679.661 uV, v_off_fin = 2677.627 uV 16:23:35 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 212.478 uVs 16:23:35 Integrated Voltage Measurement... 16:23:44 Voltage Offset Correction: v_off_ini = 2647.356 uV, v_off_fin = 2632.161 uV 16:23:44 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -220.528 uVs 16:23:44 Integrated Voltage Measurement... 16:23:53 Voltage Offset Correction: v_off_ini = 2644.083 uV, v_off_fin = 2665.126 uV 16:23:53 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 211.213 uVs 16:23:57 I1Y Average and RMS of Measurements VT = -223.079 +- 5.387 uVs I1Y = -44.616 +- 1.077 uTm 16:24:02 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:24:05 Integrated Voltage Measurement... 16:24:14 Voltage Offset Correction: v_off_ini = 2651.170 uV, v_off_fin = 2633.462 uV 16:24:14 Integrated Voltage Measurement dXYus = -1.0002 mm, dXYds = 0.0000 mm dXY = -1.0002 mm VT = -104.637 uVs 16:24:14 Integrated Voltage Measurement... 16:24:23 Voltage Offset Correction: v_off_ini = 2643.110 uV, v_off_fin = 2663.481 uV 16:24:23 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 95.977 uVs 16:24:23 Integrated Voltage Measurement... 16:24:32 Voltage Offset Correction: v_off_ini = 2652.841 uV, v_off_fin = 2654.865 uV 16:24:32 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -115.413 uVs 16:24:32 Integrated Voltage Measurement... 16:24:41 Voltage Offset Correction: v_off_ini = 2671.787 uV, v_off_fin = 2670.733 uV 16:24:41 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 91.592 uVs 16:24:41 Integrated Voltage Measurement... 16:24:50 Voltage Offset Correction: v_off_ini = 2656.939 uV, v_off_fin = 2655.196 uV 16:24:50 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -88.751 uVs 16:24:50 Integrated Voltage Measurement... 16:24:59 Voltage Offset Correction: v_off_ini = 2658.699 uV, v_off_fin = 2661.780 uV 16:24:59 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 83.493 uVs 16:25:01 I2X Average and RMS of Measurements VT = -96.644 +- 7.554 uVs I2X = 8.891 +- 0.695 uTm^2 16:25:06 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:25:09 Integrated Voltage Measurement... 16:25:18 Voltage Offset Correction: v_off_ini = 2686.137 uV, v_off_fin = 2678.305 uV 16:25:18 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -226.728 uVs 16:25:18 Integrated Voltage Measurement... 16:25:27 Voltage Offset Correction: v_off_ini = 2688.023 uV, v_off_fin = 2681.570 uV 16:25:27 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 212.706 uVs 16:25:27 Integrated Voltage Measurement... 16:25:36 Voltage Offset Correction: v_off_ini = 2673.354 uV, v_off_fin = 2673.059 uV 16:25:36 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -211.944 uVs 16:25:36 Integrated Voltage Measurement... 16:25:45 Voltage Offset Correction: v_off_ini = 2688.597 uV, v_off_fin = 2673.187 uV 16:25:45 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 202.099 uVs 16:25:45 Integrated Voltage Measurement... 16:25:54 Voltage Offset Correction: v_off_ini = 2664.629 uV, v_off_fin = 2666.647 uV 16:25:54 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -216.517 uVs 16:25:54 Integrated Voltage Measurement... 16:26:03 Voltage Offset Correction: v_off_ini = 2703.671 uV, v_off_fin = 2694.575 uV 16:26:03 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 221.811 uVs 16:26:06 I2Y Average and RMS of Measurements VT = -215.301 +- 5.859 uVs I2Y = -19.808 +- 0.539 uTm^2 16:26:16 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:26:21 Integrated Voltage Measurement... 16:26:29 Voltage Offset Correction: v_off_ini = 2683.799 uV, v_off_fin = 2684.427 uV 16:26:30 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -101.796 uVs 16:26:30 Integrated Voltage Measurement... 16:26:38 Voltage Offset Correction: v_off_ini = 2665.341 uV, v_off_fin = 2691.101 uV 16:26:39 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = 163.268 uVs 16:26:39 Integrated Voltage Measurement... 16:26:48 Voltage Offset Correction: v_off_ini = 2697.188 uV, v_off_fin = 2717.038 uV 16:26:48 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -136.261 uVs 16:26:48 Integrated Voltage Measurement... 16:26:57 Voltage Offset Correction: v_off_ini = 2720.840 uV, v_off_fin = 2721.230 uV 16:26:57 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = 138.539 uVs 16:26:57 Integrated Voltage Measurement... 16:27:06 Voltage Offset Correction: v_off_ini = 2719.349 uV, v_off_fin = 2706.402 uV 16:27:06 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -178.309 uVs 16:27:06 Integrated Voltage Measurement... 16:27:15 Voltage Offset Correction: v_off_ini = 2702.947 uV, v_off_fin = 2701.729 uV 16:27:15 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 172.492 uVs 16:27:19 I1X Average and RMS of Measurements VT = -148.444 +- 19.164 uVs I1X = 29.691 +- 3.832 uTm 16:27:21 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:27:26 Integrated Voltage Measurement... 16:27:34 Voltage Offset Correction: v_off_ini = 2694.100 uV, v_off_fin = 2695.096 uV 16:27:35 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -207.450 uVs 16:27:35 Integrated Voltage Measurement... 16:27:43 Voltage Offset Correction: v_off_ini = 2711.581 uV, v_off_fin = 2720.815 uV 16:27:44 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 214.980 uVs 16:27:44 Integrated Voltage Measurement... 16:27:52 Voltage Offset Correction: v_off_ini = 2716.406 uV, v_off_fin = 2681.944 uV 16:27:53 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -222.448 uVs 16:27:53 Integrated Voltage Measurement... 16:28:01 Voltage Offset Correction: v_off_ini = 2712.133 uV, v_off_fin = 2733.181 uV 16:28:01 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 199.321 uVs 16:28:02 Integrated Voltage Measurement... 16:28:10 Voltage Offset Correction: v_off_ini = 2711.462 uV, v_off_fin = 2703.777 uV 16:28:10 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -191.222 uVs 16:28:11 Integrated Voltage Measurement... 16:28:19 Voltage Offset Correction: v_off_ini = 2718.610 uV, v_off_fin = 2727.400 uV 16:28:19 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 209.963 uVs 16:28:24 I1Y Average and RMS of Measurements VT = -207.564 +- 4.931 uVs I1Y = -41.512 +- 0.986 uTm 16:28:29 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:28:31 Integrated Voltage Measurement... 16:28:40 Voltage Offset Correction: v_off_ini = 2723.444 uV, v_off_fin = 2707.742 uV 16:28:40 Integrated Voltage Measurement dXYus = -1.0002 mm, dXYds = 0.0000 mm dXY = -1.0002 mm VT = -102.502 uVs 16:28:40 Integrated Voltage Measurement... 16:28:49 Voltage Offset Correction: v_off_ini = 2709.372 uV, v_off_fin = 2725.703 uV 16:28:49 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 76.883 uVs 16:28:49 Integrated Voltage Measurement... 16:28:58 Voltage Offset Correction: v_off_ini = 2707.274 uV, v_off_fin = 2685.448 uV 16:28:58 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -104.400 uVs 16:28:58 Integrated Voltage Measurement... 16:29:07 Voltage Offset Correction: v_off_ini = 2698.194 uV, v_off_fin = 2714.675 uV 16:29:07 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 96.102 uVs 16:29:07 Integrated Voltage Measurement... 16:29:16 Voltage Offset Correction: v_off_ini = 2717.090 uV, v_off_fin = 2689.863 uV 16:29:16 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -103.111 uVs 16:29:16 Integrated Voltage Measurement... 16:29:25 Voltage Offset Correction: v_off_ini = 2711.087 uV, v_off_fin = 2731.710 uV 16:29:25 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 104.756 uVs 16:29:28 I2X Average and RMS of Measurements VT = -97.959 +- 6.036 uVs I2X = 9.012 +- 0.556 uTm^2 16:29:33 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:29:36 Integrated Voltage Measurement... 16:29:45 Voltage Offset Correction: v_off_ini = 2701.304 uV, v_off_fin = 2709.226 uV 16:29:45 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -222.363 uVs 16:29:45 Integrated Voltage Measurement... 16:29:54 Voltage Offset Correction: v_off_ini = 2703.336 uV, v_off_fin = 2700.164 uV 16:29:54 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = 196.765 uVs 16:29:54 Integrated Voltage Measurement... 16:30:03 Voltage Offset Correction: v_off_ini = 2705.026 uV, v_off_fin = 2736.039 uV 16:30:03 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -218.332 uVs 16:30:03 Integrated Voltage Measurement... 16:30:12 Voltage Offset Correction: v_off_ini = 2721.298 uV, v_off_fin = 2740.721 uV 16:30:12 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 191.074 uVs 16:30:12 Integrated Voltage Measurement... 16:30:21 Voltage Offset Correction: v_off_ini = 2718.803 uV, v_off_fin = 2721.499 uV 16:30:21 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -195.985 uVs 16:30:21 Integrated Voltage Measurement... 16:30:30 Voltage Offset Correction: v_off_ini = 2720.927 uV, v_off_fin = 2701.239 uV 16:30:30 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 201.103 uVs 16:30:32 I2Y Average and RMS of Measurements VT = -204.270 +- 4.509 uVs I2Y = -18.793 +- 0.415 uTm^2 16:30:43 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:30:47 Integrated Voltage Measurement... 16:30:56 Voltage Offset Correction: v_off_ini = 2737.614 uV, v_off_fin = 2725.697 uV 16:30:56 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -123.105 uVs 16:30:56 Integrated Voltage Measurement... 16:31:05 Voltage Offset Correction: v_off_ini = 2730.327 uV, v_off_fin = 2745.156 uV 16:31:05 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 160.733 uVs 16:31:05 Integrated Voltage Measurement... 16:31:14 Voltage Offset Correction: v_off_ini = 2735.263 uV, v_off_fin = 2741.146 uV 16:31:14 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -145.578 uVs 16:31:14 Integrated Voltage Measurement... 16:31:23 Voltage Offset Correction: v_off_ini = 2745.876 uV, v_off_fin = 2747.198 uV 16:31:23 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 158.626 uVs 16:31:23 Integrated Voltage Measurement... 16:31:32 Voltage Offset Correction: v_off_ini = 2743.978 uV, v_off_fin = 2749.415 uV 16:31:32 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -142.720 uVs 16:31:32 Integrated Voltage Measurement... 16:31:41 Voltage Offset Correction: v_off_ini = 2741.646 uV, v_off_fin = 2737.081 uV 16:31:41 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 163.739 uVs 16:31:45 I1X Average and RMS of Measurements VT = -149.083 +- 5.087 uVs I1X = 29.818 +- 1.018 uTm 16:31:48 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:31:52 Integrated Voltage Measurement... 16:32:01 Voltage Offset Correction: v_off_ini = 2742.050 uV, v_off_fin = 2724.402 uV 16:32:01 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -191.123 uVs 16:32:01 Integrated Voltage Measurement... 16:32:10 Voltage Offset Correction: v_off_ini = 2737.988 uV, v_off_fin = 2763.418 uV 16:32:10 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = 213.351 uVs 16:32:10 Integrated Voltage Measurement... 16:32:19 Voltage Offset Correction: v_off_ini = 2719.714 uV, v_off_fin = 2705.895 uV 16:32:19 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -193.952 uVs 16:32:19 Integrated Voltage Measurement... 16:32:28 Voltage Offset Correction: v_off_ini = 2717.790 uV, v_off_fin = 2731.288 uV 16:32:28 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 218.821 uVs 16:32:28 Integrated Voltage Measurement... 16:32:37 Voltage Offset Correction: v_off_ini = 2729.229 uV, v_off_fin = 2691.384 uV 16:32:37 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -221.112 uVs 16:32:37 Integrated Voltage Measurement... 16:32:46 Voltage Offset Correction: v_off_ini = 2695.292 uV, v_off_fin = 2739.540 uV 16:32:46 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 234.275 uVs 16:32:50 I1Y Average and RMS of Measurements VT = -212.106 +- 11.152 uVs I1Y = -42.421 +- 2.229 uTm 16:32:55 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:32:58 Integrated Voltage Measurement... 16:33:07 Voltage Offset Correction: v_off_ini = 2731.015 uV, v_off_fin = 2716.871 uV 16:33:07 Integrated Voltage Measurement dXYus = -1.0001 mm, dXYds = 0.0000 mm dXY = -1.0001 mm VT = -96.678 uVs 16:33:07 Integrated Voltage Measurement... 16:33:16 Voltage Offset Correction: v_off_ini = 2724.488 uV, v_off_fin = 2754.660 uV 16:33:16 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 108.232 uVs 16:33:16 Integrated Voltage Measurement... 16:33:25 Voltage Offset Correction: v_off_ini = 2724.626 uV, v_off_fin = 2725.260 uV 16:33:25 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -74.657 uVs 16:33:25 Integrated Voltage Measurement... 16:33:34 Voltage Offset Correction: v_off_ini = 2730.892 uV, v_off_fin = 2734.418 uV 16:33:34 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 84.079 uVs 16:33:34 Integrated Voltage Measurement... 16:33:43 Voltage Offset Correction: v_off_ini = 2710.343 uV, v_off_fin = 2718.686 uV 16:33:43 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -77.368 uVs 16:33:43 Integrated Voltage Measurement... 16:33:52 Voltage Offset Correction: v_off_ini = 2716.496 uV, v_off_fin = 2739.823 uV 16:33:52 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 104.562 uVs 16:33:54 I2X Average and RMS of Measurements VT = -90.929 +- 9.425 uVs I2X = 8.365 +- 0.867 uTm^2 16:33:59 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:34:02 Integrated Voltage Measurement... 16:34:11 Voltage Offset Correction: v_off_ini = 2732.275 uV, v_off_fin = 2737.261 uV 16:34:11 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -193.960 uVs 16:34:11 Integrated Voltage Measurement... 16:34:20 Voltage Offset Correction: v_off_ini = 2761.693 uV, v_off_fin = 2749.736 uV 16:34:20 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 211.244 uVs 16:34:20 Integrated Voltage Measurement... 16:34:29 Voltage Offset Correction: v_off_ini = 2741.454 uV, v_off_fin = 2715.147 uV 16:34:29 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -204.504 uVs 16:34:29 Integrated Voltage Measurement... 16:34:38 Voltage Offset Correction: v_off_ini = 2718.127 uV, v_off_fin = 2743.657 uV 16:34:38 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 197.546 uVs 16:34:38 Integrated Voltage Measurement... 16:34:47 Voltage Offset Correction: v_off_ini = 2742.788 uV, v_off_fin = 2750.470 uV 16:34:47 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -203.431 uVs 16:34:47 Integrated Voltage Measurement... 16:34:56 Voltage Offset Correction: v_off_ini = 2751.354 uV, v_off_fin = 2763.057 uV 16:34:56 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 191.543 uVs 16:34:59 I2Y Average and RMS of Measurements VT = -200.371 +- 2.139 uVs I2Y = -18.434 +- 0.197 uTm^2 16:35:09 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:35:14 Integrated Voltage Measurement... 16:35:23 Voltage Offset Correction: v_off_ini = 2765.423 uV, v_off_fin = 2761.551 uV 16:35:23 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -141.508 uVs 16:35:23 Integrated Voltage Measurement... 16:35:32 Voltage Offset Correction: v_off_ini = 2775.191 uV, v_off_fin = 2768.829 uV 16:35:32 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 124.328 uVs 16:35:32 Integrated Voltage Measurement... 16:35:41 Voltage Offset Correction: v_off_ini = 2745.745 uV, v_off_fin = 2760.050 uV 16:35:41 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -144.126 uVs 16:35:41 Integrated Voltage Measurement... 16:35:50 Voltage Offset Correction: v_off_ini = 2758.502 uV, v_off_fin = 2752.590 uV 16:35:50 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 190.602 uVs 16:35:50 Integrated Voltage Measurement... 16:35:59 Voltage Offset Correction: v_off_ini = 2760.713 uV, v_off_fin = 2753.365 uV 16:35:59 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -153.046 uVs 16:35:59 Integrated Voltage Measurement... 16:36:08 Voltage Offset Correction: v_off_ini = 2763.728 uV, v_off_fin = 2773.369 uV 16:36:08 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 141.739 uVs 16:36:12 I1X Average and RMS of Measurements VT = -149.225 +- 14.122 uVs I1X = 29.847 +- 2.825 uTm 16:36:14 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:36:19 Integrated Voltage Measurement... 16:36:28 Voltage Offset Correction: v_off_ini = 2780.580 uV, v_off_fin = 2752.433 uV 16:36:28 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -210.455 uVs 16:36:28 Integrated Voltage Measurement... 16:36:37 Voltage Offset Correction: v_off_ini = 2761.607 uV, v_off_fin = 2769.019 uV 16:36:37 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 218.796 uVs 16:36:37 Integrated Voltage Measurement... 16:36:46 Voltage Offset Correction: v_off_ini = 2752.456 uV, v_off_fin = 2749.275 uV 16:36:46 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -213.252 uVs 16:36:46 Integrated Voltage Measurement... 16:36:55 Voltage Offset Correction: v_off_ini = 2751.849 uV, v_off_fin = 2784.660 uV 16:36:55 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 208.525 uVs 16:36:55 Integrated Voltage Measurement... 16:37:03 Voltage Offset Correction: v_off_ini = 2775.307 uV, v_off_fin = 2737.026 uV 16:37:04 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -215.817 uVs 16:37:04 Integrated Voltage Measurement... 16:37:12 Voltage Offset Correction: v_off_ini = 2781.964 uV, v_off_fin = 2768.321 uV 16:37:12 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 208.461 uVs 16:37:17 I1Y Average and RMS of Measurements VT = -212.551 +- 1.553 uVs I1Y = -42.507 +- 0.310 uTm 16:37:22 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:37:24 Integrated Voltage Measurement... 16:37:33 Voltage Offset Correction: v_off_ini = 2755.913 uV, v_off_fin = 2738.914 uV 16:37:33 Integrated Voltage Measurement dXYus = -1.0001 mm, dXYds = 0.0000 mm dXY = -1.0001 mm VT = -92.463 uVs 16:37:33 Integrated Voltage Measurement... 16:37:42 Voltage Offset Correction: v_off_ini = 2759.256 uV, v_off_fin = 2766.744 uV 16:37:42 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 104.001 uVs 16:37:42 Integrated Voltage Measurement... 16:37:51 Voltage Offset Correction: v_off_ini = 2750.117 uV, v_off_fin = 2741.558 uV 16:37:51 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -105.239 uVs 16:37:51 Integrated Voltage Measurement... 16:38:00 Voltage Offset Correction: v_off_ini = 2750.140 uV, v_off_fin = 2753.271 uV 16:38:00 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 70.342 uVs 16:38:00 Integrated Voltage Measurement... 16:38:09 Voltage Offset Correction: v_off_ini = 2759.513 uV, v_off_fin = 2741.645 uV 16:38:09 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -120.586 uVs 16:38:10 Integrated Voltage Measurement... 16:38:18 Voltage Offset Correction: v_off_ini = 2754.820 uV, v_off_fin = 2753.426 uV 16:38:19 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 80.353 uVs 16:38:21 I2X Average and RMS of Measurements VT = -95.497 +- 5.526 uVs I2X = 8.786 +- 0.508 uTm^2 16:38:26 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:38:29 Integrated Voltage Measurement... 16:38:38 Voltage Offset Correction: v_off_ini = 2759.338 uV, v_off_fin = 2782.772 uV 16:38:38 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = -201.430 uVs 16:38:38 Integrated Voltage Measurement... 16:38:47 Voltage Offset Correction: v_off_ini = 2754.257 uV, v_off_fin = 2760.423 uV 16:38:47 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 221.902 uVs 16:38:47 Integrated Voltage Measurement... 16:38:56 Voltage Offset Correction: v_off_ini = 2763.337 uV, v_off_fin = 2769.466 uV 16:38:56 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0001 mm dXY = 1.0000 mm VT = -194.579 uVs 16:38:56 Integrated Voltage Measurement... 16:39:05 Voltage Offset Correction: v_off_ini = 2759.043 uV, v_off_fin = 2757.198 uV 16:39:05 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = -0.0001 mm dXY = -1.0000 mm VT = 215.100 uVs 16:39:05 Integrated Voltage Measurement... 16:39:14 Voltage Offset Correction: v_off_ini = 2770.894 uV, v_off_fin = 2768.701 uV 16:39:14 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -188.061 uVs 16:39:14 Integrated Voltage Measurement... 16:39:23 Voltage Offset Correction: v_off_ini = 2769.303 uV, v_off_fin = 2779.034 uV 16:39:23 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 193.629 uVs 16:39:25 I2Y Average and RMS of Measurements VT = -202.450 +- 8.666 uVs I2Y = -18.626 +- 0.798 uTm^2 16:39:36 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:39:40 Integrated Voltage Measurement... 16:39:49 Voltage Offset Correction: v_off_ini = 2769.514 uV, v_off_fin = 2755.540 uV 16:39:49 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -178.037 uVs 16:39:49 Integrated Voltage Measurement... 16:39:58 Voltage Offset Correction: v_off_ini = 2772.194 uV, v_off_fin = 2754.608 uV 16:39:58 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 196.362 uVs 16:39:58 Integrated Voltage Measurement... 16:40:07 Voltage Offset Correction: v_off_ini = 2763.347 uV, v_off_fin = 2756.075 uV 16:40:07 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -123.892 uVs 16:40:07 Integrated Voltage Measurement... 16:40:16 Voltage Offset Correction: v_off_ini = 2775.467 uV, v_off_fin = 2783.500 uV 16:40:16 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 176.100 uVs 16:40:16 Integrated Voltage Measurement... 16:40:25 Voltage Offset Correction: v_off_ini = 2773.257 uV, v_off_fin = 2765.465 uV 16:40:25 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -156.795 uVs 16:40:25 Integrated Voltage Measurement... 16:40:34 Voltage Offset Correction: v_off_ini = 2765.527 uV, v_off_fin = 2742.291 uV 16:40:34 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 175.824 uVs 16:40:38 I1X Average and RMS of Measurements VT = -167.835 +- 15.226 uVs I1X = 33.568 +- 3.046 uTm 16:40:41 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:40:45 Integrated Voltage Measurement... 16:40:54 Voltage Offset Correction: v_off_ini = 2751.845 uV, v_off_fin = 2744.093 uV 16:40:54 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -185.210 uVs 16:40:54 Integrated Voltage Measurement... 16:41:03 Voltage Offset Correction: v_off_ini = 2752.575 uV, v_off_fin = 2772.560 uV 16:41:03 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5002 mm dXY = -0.5001 mm VT = 200.028 uVs 16:41:03 Integrated Voltage Measurement... 16:41:12 Voltage Offset Correction: v_off_ini = 2752.135 uV, v_off_fin = 2733.854 uV 16:41:12 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -203.703 uVs 16:41:12 Integrated Voltage Measurement... 16:41:21 Voltage Offset Correction: v_off_ini = 2753.897 uV, v_off_fin = 2743.737 uV 16:41:21 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 188.864 uVs 16:41:21 Integrated Voltage Measurement... 16:41:30 Voltage Offset Correction: v_off_ini = 2737.622 uV, v_off_fin = 2761.190 uV 16:41:30 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -196.658 uVs 16:41:30 Integrated Voltage Measurement... 16:41:39 Voltage Offset Correction: v_off_ini = 2764.718 uV, v_off_fin = 2771.827 uV 16:41:39 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 201.057 uVs 16:41:43 I1Y Average and RMS of Measurements VT = -195.920 +- 2.560 uVs I1Y = -39.182 +- 0.514 uTm 16:41:48 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:41:51 Integrated Voltage Measurement... 16:42:00 Voltage Offset Correction: v_off_ini = 2738.804 uV, v_off_fin = 2748.312 uV 16:42:00 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -105.517 uVs 16:42:00 Integrated Voltage Measurement... 16:42:09 Voltage Offset Correction: v_off_ini = 2732.467 uV, v_off_fin = 2753.323 uV 16:42:09 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 106.428 uVs 16:42:09 Integrated Voltage Measurement... 16:42:18 Voltage Offset Correction: v_off_ini = 2745.511 uV, v_off_fin = 2750.192 uV 16:42:18 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -131.881 uVs 16:42:18 Integrated Voltage Measurement... 16:42:27 Voltage Offset Correction: v_off_ini = 2750.411 uV, v_off_fin = 2761.797 uV 16:42:27 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 125.864 uVs 16:42:27 Integrated Voltage Measurement... 16:42:36 Voltage Offset Correction: v_off_ini = 2755.910 uV, v_off_fin = 2751.847 uV 16:42:36 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -108.288 uVs 16:42:36 Integrated Voltage Measurement... 16:42:45 Voltage Offset Correction: v_off_ini = 2745.786 uV, v_off_fin = 2743.714 uV 16:42:45 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 128.148 uVs 16:42:48 I2X Average and RMS of Measurements VT = -117.688 +- 9.356 uVs I2X = 10.827 +- 0.861 uTm^2 16:42:53 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:42:55 Integrated Voltage Measurement... 16:43:04 Voltage Offset Correction: v_off_ini = 2758.188 uV, v_off_fin = 2750.187 uV 16:43:04 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -210.953 uVs 16:43:04 Integrated Voltage Measurement... 16:43:13 Voltage Offset Correction: v_off_ini = 2726.822 uV, v_off_fin = 2725.567 uV 16:43:13 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = 191.675 uVs 16:43:13 Integrated Voltage Measurement... 16:43:22 Voltage Offset Correction: v_off_ini = 2743.129 uV, v_off_fin = 2736.694 uV 16:43:22 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -184.917 uVs 16:43:22 Integrated Voltage Measurement... 16:43:31 Voltage Offset Correction: v_off_ini = 2736.158 uV, v_off_fin = 2751.712 uV 16:43:31 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 202.179 uVs 16:43:31 Integrated Voltage Measurement... 16:43:40 Voltage Offset Correction: v_off_ini = 2777.358 uV, v_off_fin = 2769.347 uV 16:43:40 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -205.390 uVs 16:43:40 Integrated Voltage Measurement... 16:43:49 Voltage Offset Correction: v_off_ini = 2757.513 uV, v_off_fin = 2753.498 uV 16:43:49 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 197.758 uVs 16:43:52 I2Y Average and RMS of Measurements VT = -198.812 +- 3.724 uVs I2Y = -18.291 +- 0.343 uTm^2 16:44:03 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:44:07 Integrated Voltage Measurement... 16:44:16 Voltage Offset Correction: v_off_ini = 2754.647 uV, v_off_fin = 2743.412 uV 16:44:16 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.4998 mm dXY = -0.4999 mm VT = -154.067 uVs 16:44:16 Integrated Voltage Measurement... 16:44:25 Voltage Offset Correction: v_off_ini = 2746.425 uV, v_off_fin = 2754.248 uV 16:44:25 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 162.544 uVs 16:44:25 Integrated Voltage Measurement... 16:44:34 Voltage Offset Correction: v_off_ini = 2754.276 uV, v_off_fin = 2738.679 uV 16:44:34 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -131.726 uVs 16:44:34 Integrated Voltage Measurement... 16:44:43 Voltage Offset Correction: v_off_ini = 2747.718 uV, v_off_fin = 2759.587 uV 16:44:43 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 142.475 uVs 16:44:43 Integrated Voltage Measurement... 16:44:52 Voltage Offset Correction: v_off_ini = 2761.716 uV, v_off_fin = 2757.643 uV 16:44:52 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -128.713 uVs 16:44:52 Integrated Voltage Measurement... 16:45:01 Voltage Offset Correction: v_off_ini = 2764.373 uV, v_off_fin = 2746.054 uV 16:45:01 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 129.329 uVs 16:45:05 I1X Average and RMS of Measurements VT = -141.476 +- 12.349 uVs I1X = 28.297 +- 2.471 uTm 16:45:08 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:45:12 Integrated Voltage Measurement... 16:45:21 Voltage Offset Correction: v_off_ini = 2754.758 uV, v_off_fin = 2737.270 uV 16:45:21 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -183.509 uVs 16:45:21 Integrated Voltage Measurement... 16:45:30 Voltage Offset Correction: v_off_ini = 2750.380 uV, v_off_fin = 2766.229 uV 16:45:30 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 195.117 uVs 16:45:30 Integrated Voltage Measurement... 16:45:39 Voltage Offset Correction: v_off_ini = 2745.964 uV, v_off_fin = 2737.215 uV 16:45:39 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -181.467 uVs 16:45:39 Integrated Voltage Measurement... 16:45:48 Voltage Offset Correction: v_off_ini = 2744.806 uV, v_off_fin = 2768.478 uV 16:45:48 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 182.586 uVs 16:45:48 Integrated Voltage Measurement... 16:45:57 Voltage Offset Correction: v_off_ini = 2766.449 uV, v_off_fin = 2752.206 uV 16:45:57 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5002 mm dXY = 0.5001 mm VT = -201.207 uVs 16:45:57 Integrated Voltage Measurement... 16:46:06 Voltage Offset Correction: v_off_ini = 2795.611 uV, v_off_fin = 2796.371 uV 16:46:06 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 177.621 uVs 16:46:10 I1Y Average and RMS of Measurements VT = -186.918 +- 3.459 uVs I1Y = -37.382 +- 0.690 uTm 16:46:15 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:46:18 Integrated Voltage Measurement... 16:46:27 Voltage Offset Correction: v_off_ini = 2785.373 uV, v_off_fin = 2780.070 uV 16:46:27 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -81.957 uVs 16:46:27 Integrated Voltage Measurement... 16:46:36 Voltage Offset Correction: v_off_ini = 2787.674 uV, v_off_fin = 2790.671 uV 16:46:36 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 113.515 uVs 16:46:36 Integrated Voltage Measurement... 16:46:45 Voltage Offset Correction: v_off_ini = 2777.891 uV, v_off_fin = 2780.615 uV 16:46:45 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -93.616 uVs 16:46:45 Integrated Voltage Measurement... 16:46:54 Voltage Offset Correction: v_off_ini = 2788.718 uV, v_off_fin = 2800.161 uV 16:46:54 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 84.131 uVs 16:46:54 Integrated Voltage Measurement... 16:47:03 Voltage Offset Correction: v_off_ini = 2798.047 uV, v_off_fin = 2780.948 uV 16:47:03 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -79.246 uVs 16:47:03 Integrated Voltage Measurement... 16:47:12 Voltage Offset Correction: v_off_ini = 2792.356 uV, v_off_fin = 2800.030 uV 16:47:12 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 114.197 uVs 16:47:14 I2X Average and RMS of Measurements VT = -94.444 +- 3.960 uVs I2X = 8.689 +- 0.364 uTm^2 16:47:19 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:47:22 Integrated Voltage Measurement... 16:47:31 Voltage Offset Correction: v_off_ini = 2802.681 uV, v_off_fin = 2828.402 uV 16:47:31 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -175.186 uVs 16:47:31 Integrated Voltage Measurement... 16:47:40 Voltage Offset Correction: v_off_ini = 2816.072 uV, v_off_fin = 2796.601 uV 16:47:40 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 194.047 uVs 16:47:40 Integrated Voltage Measurement... 16:47:49 Voltage Offset Correction: v_off_ini = 2846.029 uV, v_off_fin = 2845.947 uV 16:47:49 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -215.965 uVs 16:47:49 Integrated Voltage Measurement... 16:47:58 Voltage Offset Correction: v_off_ini = 2857.419 uV, v_off_fin = 2830.753 uV 16:47:58 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 194.760 uVs 16:47:58 Integrated Voltage Measurement... 16:48:07 Voltage Offset Correction: v_off_ini = 2831.095 uV, v_off_fin = 2822.785 uV 16:48:07 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -202.481 uVs 16:48:07 Integrated Voltage Measurement... 16:48:16 Voltage Offset Correction: v_off_ini = 2829.454 uV, v_off_fin = 2823.102 uV 16:48:16 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 200.681 uVs 16:48:18 I2Y Average and RMS of Measurements VT = -197.187 +- 9.022 uVs I2Y = -18.141 +- 0.830 uTm^2 16:48:30 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:48:35 Integrated Voltage Measurement... 16:48:43 Voltage Offset Correction: v_off_ini = 2837.407 uV, v_off_fin = 2822.086 uV 16:48:43 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -136.395 uVs 16:48:44 Integrated Voltage Measurement... 16:48:52 Voltage Offset Correction: v_off_ini = 2818.247 uV, v_off_fin = 2813.053 uV 16:48:52 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 155.694 uVs 16:48:53 Integrated Voltage Measurement... 16:49:01 Voltage Offset Correction: v_off_ini = 2818.112 uV, v_off_fin = 2804.609 uV 16:49:01 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.4999 mm dXY = -0.4999 mm VT = -167.502 uVs 16:49:02 Integrated Voltage Measurement... 16:49:10 Voltage Offset Correction: v_off_ini = 2802.641 uV, v_off_fin = 2785.137 uV 16:49:11 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = 147.788 uVs 16:49:11 Integrated Voltage Measurement... 16:49:19 Voltage Offset Correction: v_off_ini = 2804.760 uV, v_off_fin = 2815.436 uV 16:49:20 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -101.851 uVs 16:49:20 Integrated Voltage Measurement... 16:49:28 Voltage Offset Correction: v_off_ini = 2829.346 uV, v_off_fin = 2803.902 uV 16:49:29 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 124.735 uVs 16:49:33 I1X Average and RMS of Measurements VT = -138.994 +- 18.780 uVs I1X = 27.801 +- 3.757 uTm 16:49:35 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:49:39 Integrated Voltage Measurement... 16:49:48 Voltage Offset Correction: v_off_ini = 2784.422 uV, v_off_fin = 2787.322 uV 16:49:48 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -196.914 uVs 16:49:48 Integrated Voltage Measurement... 16:49:57 Voltage Offset Correction: v_off_ini = 2799.301 uV, v_off_fin = 2807.767 uV 16:49:57 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 213.676 uVs 16:49:57 Integrated Voltage Measurement... 16:50:06 Voltage Offset Correction: v_off_ini = 2794.361 uV, v_off_fin = 2785.671 uV 16:50:06 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -171.223 uVs 16:50:06 Integrated Voltage Measurement... 16:50:15 Voltage Offset Correction: v_off_ini = 2782.691 uV, v_off_fin = 2800.621 uV 16:50:15 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 183.854 uVs 16:50:15 Integrated Voltage Measurement... 16:50:24 Voltage Offset Correction: v_off_ini = 2780.199 uV, v_off_fin = 2773.695 uV 16:50:24 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -180.448 uVs 16:50:24 Integrated Voltage Measurement... 16:50:33 Voltage Offset Correction: v_off_ini = 2787.694 uV, v_off_fin = 2814.438 uV 16:50:33 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 191.509 uVs 16:50:38 I1Y Average and RMS of Measurements VT = -189.604 +- 11.618 uVs I1Y = -37.921 +- 2.324 uTm 16:50:43 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:50:45 Integrated Voltage Measurement... 16:50:54 Voltage Offset Correction: v_off_ini = 2782.249 uV, v_off_fin = 2797.444 uV 16:50:54 Integrated Voltage Measurement dXYus = -1.0001 mm, dXYds = 0.0000 mm dXY = -1.0001 mm VT = -91.912 uVs 16:50:54 Integrated Voltage Measurement... 16:51:03 Voltage Offset Correction: v_off_ini = 2797.329 uV, v_off_fin = 2789.822 uV 16:51:03 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 71.035 uVs 16:51:03 Integrated Voltage Measurement... 16:51:12 Voltage Offset Correction: v_off_ini = 2776.979 uV, v_off_fin = 2776.421 uV 16:51:12 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -92.843 uVs 16:51:12 Integrated Voltage Measurement... 16:51:21 Voltage Offset Correction: v_off_ini = 2796.798 uV, v_off_fin = 2813.562 uV 16:51:21 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 121.015 uVs 16:51:21 Integrated Voltage Measurement... 16:51:30 Voltage Offset Correction: v_off_ini = 2786.976 uV, v_off_fin = 2782.040 uV 16:51:30 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -84.538 uVs 16:51:30 Integrated Voltage Measurement... 16:51:39 Voltage Offset Correction: v_off_ini = 2804.345 uV, v_off_fin = 2799.682 uV 16:51:39 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 101.517 uVs 16:51:42 I2X Average and RMS of Measurements VT = -93.810 +- 10.407 uVs I2X = 8.630 +- 0.958 uTm^2 16:51:47 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:51:49 Integrated Voltage Measurement... 16:51:58 Voltage Offset Correction: v_off_ini = 2796.574 uV, v_off_fin = 2799.953 uV 16:51:58 Integrated Voltage Measurement dXYus = 0.9998 mm, dXYds = 0.0000 mm dXY = 0.9998 mm VT = -204.211 uVs 16:51:58 Integrated Voltage Measurement... 16:52:07 Voltage Offset Correction: v_off_ini = 2802.337 uV, v_off_fin = 2821.373 uV 16:52:07 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 202.413 uVs 16:52:08 Integrated Voltage Measurement... 16:52:16 Voltage Offset Correction: v_off_ini = 2793.945 uV, v_off_fin = 2789.323 uV 16:52:17 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -180.053 uVs 16:52:17 Integrated Voltage Measurement... 16:52:26 Voltage Offset Correction: v_off_ini = 2805.809 uV, v_off_fin = 2813.302 uV 16:52:26 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 198.613 uVs 16:52:26 Integrated Voltage Measurement... 16:52:35 Voltage Offset Correction: v_off_ini = 2818.087 uV, v_off_fin = 2803.276 uV 16:52:35 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -190.654 uVs 16:52:35 Integrated Voltage Measurement... 16:52:44 Voltage Offset Correction: v_off_ini = 2804.204 uV, v_off_fin = 2798.038 uV 16:52:44 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 192.237 uVs 16:52:46 I2Y Average and RMS of Measurements VT = -194.697 +- 6.153 uVs I2Y = -17.913 +- 0.567 uTm^2 16:52:58 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:53:02 Integrated Voltage Measurement... 16:53:11 Voltage Offset Correction: v_off_ini = 2758.602 uV, v_off_fin = 2792.684 uV 16:53:11 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.4999 mm dXY = -0.4999 mm VT = -166.856 uVs 16:53:11 Integrated Voltage Measurement... 16:53:20 Voltage Offset Correction: v_off_ini = 2787.733 uV, v_off_fin = 2797.522 uV 16:53:20 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 136.529 uVs 16:53:20 Integrated Voltage Measurement... 16:53:29 Voltage Offset Correction: v_off_ini = 2791.310 uV, v_off_fin = 2782.197 uV 16:53:29 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -120.308 uVs 16:53:29 Integrated Voltage Measurement... 16:53:38 Voltage Offset Correction: v_off_ini = 2779.170 uV, v_off_fin = 2770.286 uV 16:53:38 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 142.870 uVs 16:53:38 Integrated Voltage Measurement... 16:53:47 Voltage Offset Correction: v_off_ini = 2790.587 uV, v_off_fin = 2789.165 uV 16:53:47 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -151.235 uVs 16:53:47 Integrated Voltage Measurement... 16:53:56 Voltage Offset Correction: v_off_ini = 2789.541 uV, v_off_fin = 2793.188 uV 16:53:56 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 166.438 uVs 16:54:00 I1X Average and RMS of Measurements VT = -147.373 +- 11.536 uVs I1X = 29.476 +- 2.308 uTm 16:54:03 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:54:07 Integrated Voltage Measurement... 16:54:16 Voltage Offset Correction: v_off_ini = 2764.977 uV, v_off_fin = 2762.851 uV 16:54:16 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -178.633 uVs 16:54:16 Integrated Voltage Measurement... 16:54:25 Voltage Offset Correction: v_off_ini = 2771.308 uV, v_off_fin = 2782.726 uV 16:54:25 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 194.662 uVs 16:54:25 Integrated Voltage Measurement... 16:54:34 Voltage Offset Correction: v_off_ini = 2783.678 uV, v_off_fin = 2772.664 uV 16:54:34 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -198.394 uVs 16:54:34 Integrated Voltage Measurement... 16:54:43 Voltage Offset Correction: v_off_ini = 2748.512 uV, v_off_fin = 2760.236 uV 16:54:43 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 183.396 uVs 16:54:43 Integrated Voltage Measurement... 16:54:52 Voltage Offset Correction: v_off_ini = 2740.784 uV, v_off_fin = 2731.347 uV 16:54:52 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -194.226 uVs 16:54:52 Integrated Voltage Measurement... 16:55:01 Voltage Offset Correction: v_off_ini = 2770.153 uV, v_off_fin = 2775.305 uV 16:55:01 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 163.120 uVs 16:55:05 I1Y Average and RMS of Measurements VT = -185.405 +- 5.067 uVs I1Y = -37.081 +- 1.013 uTm 16:55:10 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:55:13 Integrated Voltage Measurement... 16:55:22 Voltage Offset Correction: v_off_ini = 2753.686 uV, v_off_fin = 2738.798 uV 16:55:22 Integrated Voltage Measurement dXYus = -1.0001 mm, dXYds = 0.0000 mm dXY = -1.0001 mm VT = -87.413 uVs 16:55:22 Integrated Voltage Measurement... 16:55:31 Voltage Offset Correction: v_off_ini = 2759.836 uV, v_off_fin = 2762.999 uV 16:55:31 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 92.555 uVs 16:55:31 Integrated Voltage Measurement... 16:55:40 Voltage Offset Correction: v_off_ini = 2761.059 uV, v_off_fin = 2739.482 uV 16:55:40 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -77.112 uVs 16:55:40 Integrated Voltage Measurement... 16:55:49 Voltage Offset Correction: v_off_ini = 2754.316 uV, v_off_fin = 2761.580 uV 16:55:49 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 105.114 uVs 16:55:49 Integrated Voltage Measurement... 16:55:58 Voltage Offset Correction: v_off_ini = 2760.491 uV, v_off_fin = 2721.400 uV 16:55:58 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -84.555 uVs 16:55:58 Integrated Voltage Measurement... 16:56:07 Voltage Offset Correction: v_off_ini = 2775.883 uV, v_off_fin = 2777.811 uV 16:56:07 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 86.758 uVs 16:56:09 I2X Average and RMS of Measurements VT = -88.918 +- 2.352 uVs I2X = 8.180 +- 0.216 uTm^2 16:56:15 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:56:17 Integrated Voltage Measurement... 16:56:26 Voltage Offset Correction: v_off_ini = 2753.064 uV, v_off_fin = 2752.855 uV 16:56:26 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -186.337 uVs 16:56:26 Integrated Voltage Measurement... 16:56:35 Voltage Offset Correction: v_off_ini = 2743.351 uV, v_off_fin = 2757.030 uV 16:56:35 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 183.837 uVs 16:56:35 Integrated Voltage Measurement... 16:56:44 Voltage Offset Correction: v_off_ini = 2744.597 uV, v_off_fin = 2753.813 uV 16:56:44 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = -185.540 uVs 16:56:44 Integrated Voltage Measurement... 16:56:53 Voltage Offset Correction: v_off_ini = 2754.462 uV, v_off_fin = 2747.663 uV 16:56:53 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 189.259 uVs 16:56:53 Integrated Voltage Measurement... 16:57:02 Voltage Offset Correction: v_off_ini = 2762.860 uV, v_off_fin = 2746.705 uV 16:57:02 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -189.422 uVs 16:57:02 Integrated Voltage Measurement... 16:57:11 Voltage Offset Correction: v_off_ini = 2747.477 uV, v_off_fin = 2743.421 uV 16:57:11 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 179.307 uVs 16:57:14 I2Y Average and RMS of Measurements VT = -185.617 +- 1.295 uVs I2Y = -17.077 +- 0.119 uTm^2 16:57:28 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:57:32 Integrated Voltage Measurement... 16:57:41 Voltage Offset Correction: v_off_ini = 2776.205 uV, v_off_fin = 2753.699 uV 16:57:41 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -136.450 uVs 16:57:41 Integrated Voltage Measurement... 16:57:50 Voltage Offset Correction: v_off_ini = 2775.763 uV, v_off_fin = 2770.991 uV 16:57:50 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 110.502 uVs 16:57:50 Integrated Voltage Measurement... 16:57:59 Voltage Offset Correction: v_off_ini = 2775.181 uV, v_off_fin = 2764.122 uV 16:57:59 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -145.173 uVs 16:57:59 Integrated Voltage Measurement... 16:58:08 Voltage Offset Correction: v_off_ini = 2756.435 uV, v_off_fin = 2745.702 uV 16:58:08 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 126.646 uVs 16:58:08 Integrated Voltage Measurement... 16:58:17 Voltage Offset Correction: v_off_ini = 2752.177 uV, v_off_fin = 2746.022 uV 16:58:17 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -128.659 uVs 16:58:17 Integrated Voltage Measurement... 16:58:26 Voltage Offset Correction: v_off_ini = 2739.178 uV, v_off_fin = 2763.901 uV 16:58:26 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 127.078 uVs 16:58:30 I1X Average and RMS of Measurements VT = -129.085 +- 5.148 uVs I1X = 25.817 +- 1.029 uTm 16:58:33 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:58:37 Integrated Voltage Measurement... 16:58:46 Voltage Offset Correction: v_off_ini = 2720.478 uV, v_off_fin = 2736.599 uV 16:58:46 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = -173.970 uVs 16:58:46 Integrated Voltage Measurement... 16:58:55 Voltage Offset Correction: v_off_ini = 2744.110 uV, v_off_fin = 2753.733 uV 16:58:55 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 206.491 uVs 16:58:55 Integrated Voltage Measurement... 16:59:04 Voltage Offset Correction: v_off_ini = 2736.569 uV, v_off_fin = 2724.041 uV 16:59:04 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -177.853 uVs 16:59:04 Integrated Voltage Measurement... 16:59:13 Voltage Offset Correction: v_off_ini = 2756.817 uV, v_off_fin = 2752.380 uV 16:59:13 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 192.341 uVs 16:59:13 Integrated Voltage Measurement... 16:59:22 Voltage Offset Correction: v_off_ini = 2763.124 uV, v_off_fin = 2739.942 uV 16:59:22 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -165.074 uVs 16:59:22 Integrated Voltage Measurement... 16:59:31 Voltage Offset Correction: v_off_ini = 2757.227 uV, v_off_fin = 2757.231 uV 16:59:31 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 201.739 uVs 16:59:35 I1Y Average and RMS of Measurements VT = -186.245 +- 2.902 uVs I1Y = -37.249 +- 0.580 uTm 16:59:40 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:59:43 Integrated Voltage Measurement... 16:59:52 Voltage Offset Correction: v_off_ini = 2736.715 uV, v_off_fin = 2747.367 uV 16:59:52 Integrated Voltage Measurement dXYus = -1.0002 mm, dXYds = 0.0000 mm dXY = -1.0002 mm VT = -88.843 uVs 16:59:52 Integrated Voltage Measurement... 17:00:01 Voltage Offset Correction: v_off_ini = 2729.088 uV, v_off_fin = 2763.994 uV 17:00:01 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 81.768 uVs 17:00:01 Integrated Voltage Measurement... 17:00:10 Voltage Offset Correction: v_off_ini = 2746.958 uV, v_off_fin = 2728.512 uV 17:00:10 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -82.166 uVs 17:00:10 Integrated Voltage Measurement... 17:00:19 Voltage Offset Correction: v_off_ini = 2736.459 uV, v_off_fin = 2737.529 uV 17:00:19 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 80.422 uVs 17:00:19 Integrated Voltage Measurement... 17:00:28 Voltage Offset Correction: v_off_ini = 2757.004 uV, v_off_fin = 2735.244 uV 17:00:28 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -82.507 uVs 17:00:28 Integrated Voltage Measurement... 17:00:37 Voltage Offset Correction: v_off_ini = 2743.358 uV, v_off_fin = 2780.444 uV 17:00:37 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 66.787 uVs 17:00:39 I2X Average and RMS of Measurements VT = -80.415 +- 4.395 uVs I2X = 7.398 +- 0.404 uTm^2 17:00:44 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 17:00:47 Integrated Voltage Measurement... 17:00:56 Voltage Offset Correction: v_off_ini = 2766.185 uV, v_off_fin = 2768.637 uV 17:00:56 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -180.671 uVs 17:00:56 Integrated Voltage Measurement... 17:01:05 Voltage Offset Correction: v_off_ini = 2766.178 uV, v_off_fin = 2770.036 uV 17:01:05 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 193.868 uVs 17:01:05 Integrated Voltage Measurement... 17:01:14 Voltage Offset Correction: v_off_ini = 2763.431 uV, v_off_fin = 2770.238 uV 17:01:14 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -201.009 uVs 17:01:14 Integrated Voltage Measurement... 17:01:23 Voltage Offset Correction: v_off_ini = 2764.810 uV, v_off_fin = 2761.318 uV 17:01:23 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 203.418 uVs 17:01:23 Integrated Voltage Measurement... 17:01:32 Voltage Offset Correction: v_off_ini = 2752.563 uV, v_off_fin = 2759.131 uV 17:01:32 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -173.318 uVs 17:01:32 Integrated Voltage Measurement... 17:01:41 Voltage Offset Correction: v_off_ini = 2783.876 uV, v_off_fin = 2789.535 uV 17:01:41 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 198.335 uVs 17:01:44 I2Y Average and RMS of Measurements VT = -191.770 +- 7.408 uVs I2Y = -17.643 +- 0.682 uTm^2 17:01:58 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 17:02:02 Integrated Voltage Measurement... 17:02:11 Voltage Offset Correction: v_off_ini = 2816.266 uV, v_off_fin = 2782.278 uV 17:02:11 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.4998 mm dXY = -0.4999 mm VT = -124.830 uVs 17:02:11 Integrated Voltage Measurement... 17:02:20 Voltage Offset Correction: v_off_ini = 2796.420 uV, v_off_fin = 2791.112 uV 17:02:20 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 110.559 uVs 17:02:20 Integrated Voltage Measurement... 17:02:29 Voltage Offset Correction: v_off_ini = 2793.530 uV, v_off_fin = 2783.350 uV 17:02:29 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -98.796 uVs 17:02:29 Integrated Voltage Measurement... 17:02:38 Voltage Offset Correction: v_off_ini = 2768.274 uV, v_off_fin = 2791.813 uV 17:02:38 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 122.940 uVs 17:02:38 Integrated Voltage Measurement... 17:02:47 Voltage Offset Correction: v_off_ini = 2787.729 uV, v_off_fin = 2790.825 uV 17:02:47 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -103.329 uVs 17:02:47 Integrated Voltage Measurement... 17:02:56 Voltage Offset Correction: v_off_ini = 2784.905 uV, v_off_fin = 2775.349 uV 17:02:56 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 109.106 uVs 17:03:00 I1X Average and RMS of Measurements VT = -111.593 +- 4.714 uVs I1X = 22.320 +- 0.944 uTm 17:03:03 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 17:03:07 Integrated Voltage Measurement... 17:03:16 Voltage Offset Correction: v_off_ini = 2776.713 uV, v_off_fin = 2778.300 uV 17:03:16 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -164.950 uVs 17:03:16 Integrated Voltage Measurement... 17:03:25 Voltage Offset Correction: v_off_ini = 2806.876 uV, v_off_fin = 2806.728 uV 17:03:25 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 195.009 uVs 17:03:25 Integrated Voltage Measurement... 17:03:34 Voltage Offset Correction: v_off_ini = 2770.350 uV, v_off_fin = 2759.694 uV 17:03:34 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -198.934 uVs 17:03:34 Integrated Voltage Measurement... 17:03:43 Voltage Offset Correction: v_off_ini = 2746.013 uV, v_off_fin = 2787.259 uV 17:03:43 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 222.171 uVs 17:03:43 Integrated Voltage Measurement... 17:03:52 Voltage Offset Correction: v_off_ini = 2739.206 uV, v_off_fin = 2777.624 uV 17:03:52 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -178.443 uVs 17:03:52 Integrated Voltage Measurement... 17:04:01 Voltage Offset Correction: v_off_ini = 2769.154 uV, v_off_fin = 2783.728 uV 17:04:01 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 196.210 uVs 17:04:05 I1Y Average and RMS of Measurements VT = -192.620 +- 13.031 uVs I1Y = -38.523 +- 2.607 uTm 17:04:10 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 17:04:13 Integrated Voltage Measurement... 17:04:22 Voltage Offset Correction: v_off_ini = 2768.085 uV, v_off_fin = 2774.264 uV 17:04:22 Integrated Voltage Measurement dXYus = -1.0001 mm, dXYds = 0.0000 mm dXY = -1.0001 mm VT = -59.514 uVs 17:04:22 Integrated Voltage Measurement... 17:04:31 Voltage Offset Correction: v_off_ini = 2784.696 uV, v_off_fin = 2787.677 uV 17:04:31 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 74.744 uVs 17:04:31 Integrated Voltage Measurement... 17:04:40 Voltage Offset Correction: v_off_ini = 2801.826 uV, v_off_fin = 2784.171 uV 17:04:40 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -67.824 uVs 17:04:40 Integrated Voltage Measurement... 17:04:49 Voltage Offset Correction: v_off_ini = 2766.579 uV, v_off_fin = 2795.451 uV 17:04:49 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 62.976 uVs 17:04:49 Integrated Voltage Measurement... 17:04:58 Voltage Offset Correction: v_off_ini = 2792.125 uV, v_off_fin = 2753.896 uV 17:04:58 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -82.275 uVs 17:04:58 Integrated Voltage Measurement... 17:05:07 Voltage Offset Correction: v_off_ini = 2794.209 uV, v_off_fin = 2787.479 uV 17:05:07 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 87.234 uVs 17:05:09 I2X Average and RMS of Measurements VT = -72.428 +- 8.745 uVs I2X = 6.663 +- 0.805 uTm^2 17:05:14 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 17:05:17 Integrated Voltage Measurement... 17:05:26 Voltage Offset Correction: v_off_ini = 2767.391 uV, v_off_fin = 2776.353 uV 17:05:26 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = -191.673 uVs 17:05:26 Integrated Voltage Measurement... 17:05:35 Voltage Offset Correction: v_off_ini = 2762.017 uV, v_off_fin = 2763.512 uV 17:05:35 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 181.183 uVs 17:05:35 Integrated Voltage Measurement... 17:05:44 Voltage Offset Correction: v_off_ini = 2774.339 uV, v_off_fin = 2753.317 uV 17:05:44 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -182.502 uVs 17:05:44 Integrated Voltage Measurement... 17:05:53 Voltage Offset Correction: v_off_ini = 2756.468 uV, v_off_fin = 2755.920 uV 17:05:53 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 161.702 uVs 17:05:53 Integrated Voltage Measurement... 17:06:02 Voltage Offset Correction: v_off_ini = 2797.011 uV, v_off_fin = 2765.506 uV 17:06:02 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -174.268 uVs 17:06:02 Integrated Voltage Measurement... 17:06:11 Voltage Offset Correction: v_off_ini = 2786.782 uV, v_off_fin = 2791.444 uV 17:06:11 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 193.499 uVs 17:06:14 I2Y Average and RMS of Measurements VT = -180.804 +- 6.241 uVs I2Y = -16.634 +- 0.574 uTm^2 17:06:28 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 17:06:32 Integrated Voltage Measurement... 17:06:41 Voltage Offset Correction: v_off_ini = 2773.878 uV, v_off_fin = 2777.269 uV 17:06:41 Integrated Voltage Measurement dXYus = -0.4998 mm, dXYds = -0.4999 mm dXY = -0.4999 mm VT = -97.182 uVs 17:06:41 Integrated Voltage Measurement... 17:06:50 Voltage Offset Correction: v_off_ini = 2788.818 uV, v_off_fin = 2790.584 uV 17:06:50 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 99.449 uVs 17:06:50 Integrated Voltage Measurement... 17:06:59 Voltage Offset Correction: v_off_ini = 2772.791 uV, v_off_fin = 2768.296 uV 17:06:59 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -111.385 uVs 17:06:59 Integrated Voltage Measurement... 17:07:08 Voltage Offset Correction: v_off_ini = 2780.786 uV, v_off_fin = 2761.743 uV 17:07:08 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 95.285 uVs 17:07:08 Integrated Voltage Measurement... 17:07:17 Voltage Offset Correction: v_off_ini = 2808.354 uV, v_off_fin = 2796.796 uV 17:07:17 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -114.748 uVs 17:07:17 Integrated Voltage Measurement... 17:07:26 Voltage Offset Correction: v_off_ini = 2808.668 uV, v_off_fin = 2812.505 uV 17:07:26 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 109.320 uVs 17:07:30 I1X Average and RMS of Measurements VT = -104.561 +- 5.667 uVs I1X = 20.914 +- 1.132 uTm 17:07:33 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 17:07:37 Integrated Voltage Measurement... 17:07:46 Voltage Offset Correction: v_off_ini = 2809.454 uV, v_off_fin = 2812.773 uV 17:07:46 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = -213.693 uVs 17:07:46 Integrated Voltage Measurement... 17:07:55 Voltage Offset Correction: v_off_ini = 2833.836 uV, v_off_fin = 2831.597 uV 17:07:55 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 184.837 uVs 17:07:55 Integrated Voltage Measurement... 17:08:04 Voltage Offset Correction: v_off_ini = 2803.039 uV, v_off_fin = 2781.724 uV 17:08:04 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -202.285 uVs 17:08:04 Integrated Voltage Measurement... 17:08:13 Voltage Offset Correction: v_off_ini = 2793.782 uV, v_off_fin = 2815.652 uV 17:08:13 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 197.451 uVs 17:08:13 Integrated Voltage Measurement... 17:08:22 Voltage Offset Correction: v_off_ini = 2773.100 uV, v_off_fin = 2763.867 uV 17:08:22 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -179.408 uVs 17:08:22 Integrated Voltage Measurement... 17:08:31 Voltage Offset Correction: v_off_ini = 2758.750 uV, v_off_fin = 2784.426 uV 17:08:31 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 224.975 uVs 17:08:35 I1Y Average and RMS of Measurements VT = -200.441 +- 1.262 uVs I1Y = -40.086 +- 0.251 uTm 17:08:40 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 17:08:43 Integrated Voltage Measurement... 17:08:52 Voltage Offset Correction: v_off_ini = 2789.661 uV, v_off_fin = 2789.870 uV 17:08:52 Integrated Voltage Measurement dXYus = -1.0001 mm, dXYds = 0.0000 mm dXY = -1.0001 mm VT = -55.401 uVs 17:08:52 Integrated Voltage Measurement... 17:09:01 Voltage Offset Correction: v_off_ini = 2804.798 uV, v_off_fin = 2808.377 uV 17:09:01 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 74.767 uVs 17:09:01 Integrated Voltage Measurement... 17:09:10 Voltage Offset Correction: v_off_ini = 2824.039 uV, v_off_fin = 2822.866 uV 17:09:10 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -54.728 uVs 17:09:10 Integrated Voltage Measurement... 17:09:19 Voltage Offset Correction: v_off_ini = 2837.056 uV, v_off_fin = 2829.535 uV 17:09:19 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 49.030 uVs 17:09:19 Integrated Voltage Measurement... 17:09:28 Voltage Offset Correction: v_off_ini = 2809.790 uV, v_off_fin = 2809.205 uV 17:09:28 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -57.288 uVs 17:09:28 Integrated Voltage Measurement... 17:09:37 Voltage Offset Correction: v_off_ini = 2804.738 uV, v_off_fin = 2798.615 uV 17:09:37 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 48.035 uVs 17:09:39 I2X Average and RMS of Measurements VT = -56.541 +- 6.049 uVs I2X = 5.202 +- 0.556 uTm^2 17:09:44 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 17:09:47 Integrated Voltage Measurement... 17:09:56 Voltage Offset Correction: v_off_ini = 2814.650 uV, v_off_fin = 2796.415 uV 17:09:56 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = -206.095 uVs 17:09:56 Integrated Voltage Measurement... 17:10:05 Voltage Offset Correction: v_off_ini = 2799.251 uV, v_off_fin = 2789.084 uV 17:10:05 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 203.958 uVs 17:10:05 Integrated Voltage Measurement... 17:10:14 Voltage Offset Correction: v_off_ini = 2802.241 uV, v_off_fin = 2796.844 uV 17:10:14 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -217.118 uVs 17:10:14 Integrated Voltage Measurement... 17:10:23 Voltage Offset Correction: v_off_ini = 2802.710 uV, v_off_fin = 2807.333 uV 17:10:23 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 206.480 uVs 17:10:23 Integrated Voltage Measurement... 17:10:32 Voltage Offset Correction: v_off_ini = 2788.345 uV, v_off_fin = 2791.055 uV 17:10:32 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -185.536 uVs 17:10:32 Integrated Voltage Measurement... 17:10:41 Voltage Offset Correction: v_off_ini = 2794.791 uV, v_off_fin = 2813.000 uV 17:10:41 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 204.142 uVs 17:10:43 I2Y Average and RMS of Measurements VT = -203.888 +- 6.970 uVs I2Y = -18.758 +- 0.641 uTm^2 17:10:57 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 17:11:02 Integrated Voltage Measurement... 17:11:10 Voltage Offset Correction: v_off_ini = 2782.271 uV, v_off_fin = 2759.174 uV 17:11:11 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.4999 mm dXY = -0.4999 mm VT = -97.331 uVs 17:11:11 Integrated Voltage Measurement... 17:11:19 Voltage Offset Correction: v_off_ini = 2772.495 uV, v_off_fin = 2767.429 uV 17:11:19 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 83.420 uVs 17:11:20 Integrated Voltage Measurement... 17:11:28 Voltage Offset Correction: v_off_ini = 2750.453 uV, v_off_fin = 2757.162 uV 17:11:28 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -108.417 uVs 17:11:29 Integrated Voltage Measurement... 17:11:37 Voltage Offset Correction: v_off_ini = 2767.511 uV, v_off_fin = 2748.787 uV 17:11:37 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 84.179 uVs 17:11:38 Integrated Voltage Measurement... 17:11:46 Voltage Offset Correction: v_off_ini = 2774.741 uV, v_off_fin = 2752.611 uV 17:11:46 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -108.280 uVs 17:11:46 Integrated Voltage Measurement... 17:11:55 Voltage Offset Correction: v_off_ini = 2769.325 uV, v_off_fin = 2778.691 uV 17:11:55 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 97.652 uVs 17:12:00 I1X Average and RMS of Measurements VT = -96.547 +- 5.143 uVs I1X = 19.310 +- 1.028 uTm 17:12:02 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 17:12:06 Integrated Voltage Measurement... 17:12:15 Voltage Offset Correction: v_off_ini = 2778.228 uV, v_off_fin = 2781.752 uV 17:12:15 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -181.701 uVs 17:12:15 Integrated Voltage Measurement... 17:12:24 Voltage Offset Correction: v_off_ini = 2790.870 uV, v_off_fin = 2802.366 uV 17:12:24 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 218.065 uVs 17:12:24 Integrated Voltage Measurement... 17:12:33 Voltage Offset Correction: v_off_ini = 2819.937 uV, v_off_fin = 2792.905 uV 17:12:33 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -199.009 uVs 17:12:33 Integrated Voltage Measurement... 17:12:42 Voltage Offset Correction: v_off_ini = 2775.332 uV, v_off_fin = 2798.097 uV 17:12:42 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 203.044 uVs 17:12:42 Integrated Voltage Measurement... 17:12:51 Voltage Offset Correction: v_off_ini = 2785.264 uV, v_off_fin = 2769.491 uV 17:12:51 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -204.817 uVs 17:12:51 Integrated Voltage Measurement... 17:13:00 Voltage Offset Correction: v_off_ini = 2765.235 uV, v_off_fin = 2798.974 uV 17:13:00 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 228.898 uVs 17:13:04 I1Y Average and RMS of Measurements VT = -205.922 +- 7.746 uVs I1Y = -41.180 +- 1.549 uTm 17:13:09 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 17:13:12 Integrated Voltage Measurement... 17:13:21 Voltage Offset Correction: v_off_ini = 2773.365 uV, v_off_fin = 2771.882 uV 17:13:21 Integrated Voltage Measurement dXYus = -1.0001 mm, dXYds = 0.0000 mm dXY = -1.0001 mm VT = -34.832 uVs 17:13:21 Integrated Voltage Measurement... 17:13:30 Voltage Offset Correction: v_off_ini = 2774.824 uV, v_off_fin = 2777.715 uV 17:13:30 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 58.371 uVs 17:13:30 Integrated Voltage Measurement... 17:13:39 Voltage Offset Correction: v_off_ini = 2784.511 uV, v_off_fin = 2784.544 uV 17:13:39 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -34.792 uVs 17:13:39 Integrated Voltage Measurement... 17:13:48 Voltage Offset Correction: v_off_ini = 2797.514 uV, v_off_fin = 2803.358 uV 17:13:48 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 54.532 uVs 17:13:48 Integrated Voltage Measurement... 17:13:57 Voltage Offset Correction: v_off_ini = 2787.970 uV, v_off_fin = 2776.617 uV 17:13:57 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -44.003 uVs 17:13:57 Integrated Voltage Measurement... 17:14:06 Voltage Offset Correction: v_off_ini = 2794.639 uV, v_off_fin = 2779.564 uV 17:14:06 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 37.058 uVs 17:14:08 I2X Average and RMS of Measurements VT = -43.931 +- 2.532 uVs I2X = 4.042 +- 0.233 uTm^2 17:14:14 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 17:14:16 Integrated Voltage Measurement... 17:14:25 Voltage Offset Correction: v_off_ini = 2792.987 uV, v_off_fin = 2782.961 uV 17:14:25 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -172.714 uVs 17:14:25 Integrated Voltage Measurement... 17:14:34 Voltage Offset Correction: v_off_ini = 2793.534 uV, v_off_fin = 2816.441 uV 17:14:34 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 189.770 uVs 17:14:34 Integrated Voltage Measurement... 17:14:43 Voltage Offset Correction: v_off_ini = 2814.917 uV, v_off_fin = 2808.156 uV 17:14:43 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -182.071 uVs 17:14:43 Integrated Voltage Measurement... 17:14:52 Voltage Offset Correction: v_off_ini = 2799.755 uV, v_off_fin = 2812.613 uV 17:14:52 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 207.380 uVs 17:14:52 Integrated Voltage Measurement... 17:15:01 Voltage Offset Correction: v_off_ini = 2809.758 uV, v_off_fin = 2831.450 uV 17:15:01 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -185.928 uVs 17:15:01 Integrated Voltage Measurement... 17:15:10 Voltage Offset Correction: v_off_ini = 2824.707 uV, v_off_fin = 2835.915 uV 17:15:10 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 202.056 uVs 17:15:13 I2Y Average and RMS of Measurements VT = -189.986 +- 6.191 uVs I2Y = -17.479 +- 0.570 uTm^2 17:15:27 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 17:15:31 Integrated Voltage Measurement... 17:15:40 Voltage Offset Correction: v_off_ini = 2815.187 uV, v_off_fin = 2809.544 uV 17:15:40 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.4999 mm dXY = -0.4999 mm VT = -66.205 uVs 17:15:40 Integrated Voltage Measurement... 17:15:49 Voltage Offset Correction: v_off_ini = 2824.436 uV, v_off_fin = 2831.804 uV 17:15:49 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 82.005 uVs 17:15:49 Integrated Voltage Measurement... 17:15:58 Voltage Offset Correction: v_off_ini = 2821.069 uV, v_off_fin = 2815.650 uV 17:15:58 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -96.725 uVs 17:15:58 Integrated Voltage Measurement... 17:16:07 Voltage Offset Correction: v_off_ini = 2812.637 uV, v_off_fin = 2820.001 uV 17:16:07 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 58.337 uVs 17:16:07 Integrated Voltage Measurement... 17:16:16 Voltage Offset Correction: v_off_ini = 2814.346 uV, v_off_fin = 2826.227 uV 17:16:16 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -68.523 uVs 17:16:16 Integrated Voltage Measurement... 17:16:25 Voltage Offset Correction: v_off_ini = 2825.761 uV, v_off_fin = 2823.932 uV 17:16:25 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 72.640 uVs 17:16:29 I1X Average and RMS of Measurements VT = -74.072 +- 2.837 uVs I1X = 14.815 +- 0.567 uTm 17:16:32 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 17:16:36 Integrated Voltage Measurement... 17:16:45 Voltage Offset Correction: v_off_ini = 2816.008 uV, v_off_fin = 2812.066 uV 17:16:45 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -185.453 uVs 17:16:45 Integrated Voltage Measurement... 17:16:54 Voltage Offset Correction: v_off_ini = 2805.800 uV, v_off_fin = 2828.148 uV 17:16:54 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 217.959 uVs 17:16:54 Integrated Voltage Measurement... 17:17:03 Voltage Offset Correction: v_off_ini = 2806.808 uV, v_off_fin = 2795.319 uV 17:17:03 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -200.790 uVs 17:17:03 Integrated Voltage Measurement... 17:17:12 Voltage Offset Correction: v_off_ini = 2801.352 uV, v_off_fin = 2802.966 uV 17:17:12 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 207.335 uVs 17:17:12 Integrated Voltage Measurement... 17:17:21 Voltage Offset Correction: v_off_ini = 2801.375 uV, v_off_fin = 2801.282 uV 17:17:21 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -181.572 uVs 17:17:21 Integrated Voltage Measurement... 17:17:30 Voltage Offset Correction: v_off_ini = 2824.631 uV, v_off_fin = 2830.351 uV 17:17:30 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 207.005 uVs 17:17:34 I1Y Average and RMS of Measurements VT = -200.019 +- 4.165 uVs I1Y = -40.000 +- 0.833 uTm 17:17:39 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 17:17:42 Integrated Voltage Measurement... 17:17:51 Voltage Offset Correction: v_off_ini = 2798.976 uV, v_off_fin = 2800.892 uV 17:17:51 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -37.223 uVs 17:17:51 Integrated Voltage Measurement... 17:18:00 Voltage Offset Correction: v_off_ini = 2805.082 uV, v_off_fin = 2825.459 uV 17:18:00 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 55.554 uVs 17:18:00 Integrated Voltage Measurement... 17:18:09 Voltage Offset Correction: v_off_ini = 2789.979 uV, v_off_fin = 2791.635 uV 17:18:09 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -58.838 uVs 17:18:09 Integrated Voltage Measurement... 17:18:18 Voltage Offset Correction: v_off_ini = 2797.579 uV, v_off_fin = 2799.097 uV 17:18:18 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 43.973 uVs 17:18:18 Integrated Voltage Measurement... 17:18:27 Voltage Offset Correction: v_off_ini = 2780.520 uV, v_off_fin = 2785.424 uV 17:18:27 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -10.663 uVs 17:18:27 Integrated Voltage Measurement... 17:18:36 Voltage Offset Correction: v_off_ini = 2798.329 uV, v_off_fin = 2817.312 uV 17:18:36 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 44.098 uVs 17:18:38 I2X Average and RMS of Measurements VT = -41.725 +- 10.348 uVs I2X = 3.839 +- 0.952 uTm^2 17:18:43 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 17:18:46 Integrated Voltage Measurement... 17:18:55 Voltage Offset Correction: v_off_ini = 2780.854 uV, v_off_fin = 2779.841 uV 17:18:55 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -220.605 uVs 17:18:55 Integrated Voltage Measurement... 17:19:04 Voltage Offset Correction: v_off_ini = 2782.350 uV, v_off_fin = 2791.757 uV 17:19:04 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 206.684 uVs 17:19:04 Integrated Voltage Measurement... 17:19:13 Voltage Offset Correction: v_off_ini = 2781.601 uV, v_off_fin = 2782.474 uV 17:19:13 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -187.999 uVs 17:19:13 Integrated Voltage Measurement... 17:19:22 Voltage Offset Correction: v_off_ini = 2804.834 uV, v_off_fin = 2804.944 uV 17:19:22 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 200.849 uVs 17:19:22 Integrated Voltage Measurement... 17:19:31 Voltage Offset Correction: v_off_ini = 2810.512 uV, v_off_fin = 2812.200 uV 17:19:31 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -214.643 uVs 17:19:31 Integrated Voltage Measurement... 17:19:40 Voltage Offset Correction: v_off_ini = 2819.865 uV, v_off_fin = 2840.686 uV 17:19:40 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 183.626 uVs 17:19:42 I2Y Average and RMS of Measurements VT = -202.401 +- 8.180 uVs I2Y = -18.621 +- 0.753 uTm^2 17:19:56 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 17:20:01 Integrated Voltage Measurement... 17:20:10 Voltage Offset Correction: v_off_ini = 2835.820 uV, v_off_fin = 2837.059 uV 17:20:10 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.4998 mm dXY = -0.4999 mm VT = -80.279 uVs 17:20:10 Integrated Voltage Measurement... 17:20:19 Voltage Offset Correction: v_off_ini = 2822.408 uV, v_off_fin = 2808.397 uV 17:20:19 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 62.555 uVs 17:20:19 Integrated Voltage Measurement... 17:20:28 Voltage Offset Correction: v_off_ini = 2833.683 uV, v_off_fin = 2813.498 uV 17:20:28 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -87.862 uVs 17:20:28 Integrated Voltage Measurement... 17:20:37 Voltage Offset Correction: v_off_ini = 2808.163 uV, v_off_fin = 2823.466 uV 17:20:37 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 79.838 uVs 17:20:37 Integrated Voltage Measurement... 17:20:46 Voltage Offset Correction: v_off_ini = 2820.075 uV, v_off_fin = 2808.557 uV 17:20:46 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -65.176 uVs 17:20:46 Integrated Voltage Measurement... 17:20:55 Voltage Offset Correction: v_off_ini = 2838.511 uV, v_off_fin = 2846.702 uV 17:20:55 Integrated Voltage Measurement dXYus = 0.4998 mm, dXYds = 0.5000 mm dXY = 0.4999 mm VT = 94.647 uVs 17:20:59 I1X Average and RMS of Measurements VT = -78.393 +- 5.188 uVs I1X = 15.681 +- 1.037 uTm 17:21:01 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 17:21:05 Integrated Voltage Measurement... 17:21:14 Voltage Offset Correction: v_off_ini = 2812.858 uV, v_off_fin = 2805.126 uV 17:21:14 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = -201.714 uVs 17:21:14 Integrated Voltage Measurement... 17:21:23 Voltage Offset Correction: v_off_ini = 2813.097 uV, v_off_fin = 2831.555 uV 17:21:23 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 207.946 uVs 17:21:23 Integrated Voltage Measurement... 17:21:32 Voltage Offset Correction: v_off_ini = 2805.067 uV, v_off_fin = 2791.314 uV 17:21:32 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -199.195 uVs 17:21:32 Integrated Voltage Measurement... 17:21:41 Voltage Offset Correction: v_off_ini = 2809.191 uV, v_off_fin = 2837.456 uV 17:21:41 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 213.167 uVs 17:21:41 Integrated Voltage Measurement... 17:21:50 Voltage Offset Correction: v_off_ini = 2835.087 uV, v_off_fin = 2812.460 uV 17:21:50 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -218.297 uVs 17:21:50 Integrated Voltage Measurement... 17:21:59 Voltage Offset Correction: v_off_ini = 2807.473 uV, v_off_fin = 2805.774 uV 17:21:59 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 195.067 uVs 17:22:04 I1Y Average and RMS of Measurements VT = -205.898 +- 0.782 uVs I1Y = -41.180 +- 0.156 uTm 17:22:09 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 17:22:11 Integrated Voltage Measurement... 17:22:20 Voltage Offset Correction: v_off_ini = 2764.918 uV, v_off_fin = 2769.101 uV 17:22:20 Integrated Voltage Measurement dXYus = -1.0001 mm, dXYds = 0.0000 mm dXY = -1.0001 mm VT = -6.192 uVs 17:22:20 Integrated Voltage Measurement... 17:22:29 Voltage Offset Correction: v_off_ini = 2798.163 uV, v_off_fin = 2811.750 uV 17:22:29 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 62.459 uVs 17:22:29 Integrated Voltage Measurement... 17:22:38 Voltage Offset Correction: v_off_ini = 2798.063 uV, v_off_fin = 2788.843 uV 17:22:38 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -48.450 uVs 17:22:38 Integrated Voltage Measurement... 17:22:47 Voltage Offset Correction: v_off_ini = 2804.235 uV, v_off_fin = 2808.935 uV 17:22:47 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 69.526 uVs 17:22:47 Integrated Voltage Measurement... 17:22:56 Voltage Offset Correction: v_off_ini = 2802.856 uV, v_off_fin = 2814.322 uV 17:22:56 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -31.022 uVs 17:22:56 Integrated Voltage Measurement... 17:23:05 Voltage Offset Correction: v_off_ini = 2815.904 uV, v_off_fin = 2835.596 uV 17:23:05 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 50.616 uVs 17:23:08 I2X Average and RMS of Measurements VT = -44.711 +- 10.438 uVs I2X = 4.113 +- 0.960 uTm^2 17:23:13 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 17:23:15 Integrated Voltage Measurement... 17:23:24 Voltage Offset Correction: v_off_ini = 2830.105 uV, v_off_fin = 2841.185 uV 17:23:24 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -181.036 uVs 17:23:24 Integrated Voltage Measurement... 17:23:33 Voltage Offset Correction: v_off_ini = 2819.861 uV, v_off_fin = 2815.577 uV 17:23:33 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = 219.484 uVs 17:23:33 Integrated Voltage Measurement... 17:23:42 Voltage Offset Correction: v_off_ini = 2787.376 uV, v_off_fin = 2796.376 uV 17:23:42 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -220.293 uVs 17:23:42 Integrated Voltage Measurement... 17:23:51 Voltage Offset Correction: v_off_ini = 2814.326 uV, v_off_fin = 2821.707 uV 17:23:51 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 206.572 uVs 17:23:51 Integrated Voltage Measurement... 17:24:00 Voltage Offset Correction: v_off_ini = 2825.250 uV, v_off_fin = 2797.523 uV 17:24:00 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -212.730 uVs 17:24:00 Integrated Voltage Measurement... 17:24:09 Voltage Offset Correction: v_off_ini = 2810.657 uV, v_off_fin = 2815.736 uV 17:24:09 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 213.021 uVs 17:24:12 I2Y Average and RMS of Measurements VT = -208.856 +- 6.083 uVs I2Y = -19.215 +- 0.559 uTm^2 17:24:14 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 08:31:03 Wire Position: X0 = 50.0000 mm, Y0 = 0.0000 mm