The phase shifter was re-tuned before the final measurements to meet the field integrals tolerances. The change in field integrals happened between DS001 and DS002. We didn’t measure off-axis field integrals for DS002-DS005, but there is on-axis change of a few G-cm between DS001 and DS006. It was enough to bring off-axis integrals out of tolerance. The phase integral was checked at 15mm gap before re-tuning. DS006, 15mm gap, PI=6.8496 DS005, 15mm gap, PI=6.8494 DS001, 15mm gap, PI=6.8510 The change in PI (DS006-DS001) is small (2*10-4). DS006, 15mm gap, PI=6.828 (after tuning).