SLAC Magnetic Measurements Date: 08-25-2026 Time: 10:15:08 Project: LCLS-II-HE Magnet Type: Phase Shifter Magnet Name: HE_SXPS_15081 Dataset Number: 6 Run Purpose: Final Results Run Type: Run Number: 8 Measurement Device: Stretched Wire Operator: YL Comment: On Axis 10:15:15 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:15:17 Motion Parameters, X Controller, Axis 1 acc = 0.001000 m/s/s dcl = 0.001000 m/s/s vel = 0.001000 m/s pos = 0.000000 m kp = 188.600000 ki = 820.000000 kd = 491.400000 db = 0.000000 10:15:19 Motion Parameters, X Controller, Axis 2 acc = 0.001000 m/s/s dcl = 0.001000 m/s/s vel = 0.001000 m/s pos = 0.000000 m kp = 188.600000 ki = 820.000000 kd = 491.400000 db = 0.000000 10:15:21 Motion Parameters, Y Controller, Axis 1 acc = 0.001000 m/s/s dcl = 0.001000 m/s/s vel = 0.001000 m/s pos = 0.000000 m kp = 188.600000 ki = 820.000000 kd = 491.400000 db = 0.000000 10:15:22 Motion Parameters, Y Controller, Axis 2 acc = 0.001000 m/s/s dcl = 0.001000 m/s/s vel = 0.001000 m/s pos = 0.000000 m kp = 188.600000 ki = 820.000000 kd = 491.400000 db = 0.000000 10:15:57 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:16:01 Integrated Voltage Measurement... 10:16:10 Voltage Offset Correction: v_off_ini = 2083.197 uV, v_off_fin = 2109.958 uV 10:16:10 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -128.410 uVs 10:16:10 Integrated Voltage Measurement... 10:16:19 Voltage Offset Correction: v_off_ini = 2109.614 uV, v_off_fin = 2114.873 uV 10:16:19 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 128.790 uVs 10:16:19 Integrated Voltage Measurement... 10:16:28 Voltage Offset Correction: v_off_ini = 2103.382 uV, v_off_fin = 2124.295 uV 10:16:28 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -96.893 uVs 10:16:28 Integrated Voltage Measurement... 10:16:37 Voltage Offset Correction: v_off_ini = 2123.776 uV, v_off_fin = 2112.580 uV 10:16:37 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 122.610 uVs 10:16:37 Integrated Voltage Measurement... 10:16:46 Voltage Offset Correction: v_off_ini = 2109.861 uV, v_off_fin = 2110.866 uV 10:16:46 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -134.652 uVs 10:16:46 Integrated Voltage Measurement... 10:16:55 Voltage Offset Correction: v_off_ini = 2113.134 uV, v_off_fin = 2114.408 uV 10:16:55 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 143.079 uVs 10:17:00 I1X Average and RMS of Measurements VT = -125.739 +- 12.057 uVs I1X = 25.148 +- 2.411 uTm 10:17:02 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:17:06 Integrated Voltage Measurement... 10:17:15 Voltage Offset Correction: v_off_ini = 2101.160 uV, v_off_fin = 2093.660 uV 10:17:15 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -206.930 uVs 10:17:15 Integrated Voltage Measurement... 10:17:24 Voltage Offset Correction: v_off_ini = 2119.754 uV, v_off_fin = 2123.641 uV 10:17:24 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 228.549 uVs 10:17:24 Integrated Voltage Measurement... 10:17:33 Voltage Offset Correction: v_off_ini = 2095.276 uV, v_off_fin = 2080.713 uV 10:17:33 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -227.510 uVs 10:17:33 Integrated Voltage Measurement... 10:17:42 Voltage Offset Correction: v_off_ini = 2096.379 uV, v_off_fin = 2118.685 uV 10:17:42 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 230.562 uVs 10:17:42 Integrated Voltage Measurement... 10:17:51 Voltage Offset Correction: v_off_ini = 2097.412 uV, v_off_fin = 2086.611 uV 10:17:51 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -222.330 uVs 10:17:51 Integrated Voltage Measurement... 10:18:00 Voltage Offset Correction: v_off_ini = 2102.557 uV, v_off_fin = 2101.120 uV 10:18:00 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 246.274 uVs 10:18:04 I1Y Average and RMS of Measurements VT = -227.026 +- 6.910 uVs I1Y = -45.401 +- 1.382 uTm 10:18:09 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:18:12 Integrated Voltage Measurement... 10:18:21 Voltage Offset Correction: v_off_ini = 2109.956 uV, v_off_fin = 2098.015 uV 10:18:21 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -100.367 uVs 10:18:21 Integrated Voltage Measurement... 10:18:30 Voltage Offset Correction: v_off_ini = 2121.968 uV, v_off_fin = 2123.758 uV 10:18:30 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 118.242 uVs 10:18:30 Integrated Voltage Measurement... 10:18:39 Voltage Offset Correction: v_off_ini = 2103.917 uV, v_off_fin = 2118.485 uV 10:18:39 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -102.865 uVs 10:18:39 Integrated Voltage Measurement... 10:18:48 Voltage Offset Correction: v_off_ini = 2114.208 uV, v_off_fin = 2115.441 uV 10:18:48 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 98.085 uVs 10:18:48 Integrated Voltage Measurement... 10:18:57 Voltage Offset Correction: v_off_ini = 2126.607 uV, v_off_fin = 2120.645 uV 10:18:57 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -120.058 uVs 10:18:57 Integrated Voltage Measurement... 10:19:06 Voltage Offset Correction: v_off_ini = 2129.602 uV, v_off_fin = 2128.033 uV 10:19:06 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 81.992 uVs 10:19:09 I2X Average and RMS of Measurements VT = -103.602 +- 4.039 uVs I2X = 9.531 +- 0.372 uTm^2 10:19:14 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:19:16 Integrated Voltage Measurement... 10:19:25 Voltage Offset Correction: v_off_ini = 2164.269 uV, v_off_fin = 2147.795 uV 10:19:25 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -233.505 uVs 10:19:25 Integrated Voltage Measurement... 10:19:34 Voltage Offset Correction: v_off_ini = 2148.550 uV, v_off_fin = 2142.589 uV 10:19:34 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 228.680 uVs 10:19:34 Integrated Voltage Measurement... 10:19:43 Voltage Offset Correction: v_off_ini = 2163.693 uV, v_off_fin = 2162.189 uV 10:19:43 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -249.515 uVs 10:19:43 Integrated Voltage Measurement... 10:19:52 Voltage Offset Correction: v_off_ini = 2133.594 uV, v_off_fin = 2149.355 uV 10:19:52 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 224.394 uVs 10:19:52 Integrated Voltage Measurement... 10:20:01 Voltage Offset Correction: v_off_ini = 2150.236 uV, v_off_fin = 2159.569 uV 10:20:01 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -237.227 uVs 10:20:01 Integrated Voltage Measurement... 10:20:10 Voltage Offset Correction: v_off_ini = 2184.241 uV, v_off_fin = 2166.676 uV 10:20:10 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 208.612 uVs 10:20:13 I2Y Average and RMS of Measurements VT = -230.322 +- 5.755 uVs I2Y = -21.190 +- 0.530 uTm^2 10:20:23 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:20:28 Integrated Voltage Measurement... 10:20:37 Voltage Offset Correction: v_off_ini = 2181.556 uV, v_off_fin = 2179.853 uV 10:20:37 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -132.730 uVs 10:20:37 Integrated Voltage Measurement... 10:20:46 Voltage Offset Correction: v_off_ini = 2179.635 uV, v_off_fin = 2200.767 uV 10:20:46 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 121.655 uVs 10:20:46 Integrated Voltage Measurement... 10:20:55 Voltage Offset Correction: v_off_ini = 2173.225 uV, v_off_fin = 2158.537 uV 10:20:55 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -111.382 uVs 10:20:55 Integrated Voltage Measurement... 10:21:04 Voltage Offset Correction: v_off_ini = 2155.353 uV, v_off_fin = 2170.453 uV 10:21:04 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 141.469 uVs 10:21:04 Integrated Voltage Measurement... 10:21:13 Voltage Offset Correction: v_off_ini = 2151.135 uV, v_off_fin = 2144.581 uV 10:21:13 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -96.730 uVs 10:21:13 Integrated Voltage Measurement... 10:21:22 Voltage Offset Correction: v_off_ini = 2163.384 uV, v_off_fin = 2175.728 uV 10:21:22 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 97.604 uVs 10:21:26 I1X Average and RMS of Measurements VT = -116.929 +- 13.977 uVs I1X = 23.386 +- 2.795 uTm 10:21:28 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:21:33 Integrated Voltage Measurement... 10:21:42 Voltage Offset Correction: v_off_ini = 2165.322 uV, v_off_fin = 2169.596 uV 10:21:42 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -208.689 uVs 10:21:42 Integrated Voltage Measurement... 10:21:51 Voltage Offset Correction: v_off_ini = 2156.292 uV, v_off_fin = 2181.711 uV 10:21:51 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 221.255 uVs 10:21:51 Integrated Voltage Measurement... 10:21:59 Voltage Offset Correction: v_off_ini = 2171.242 uV, v_off_fin = 2164.026 uV 10:22:00 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -191.878 uVs 10:22:00 Integrated Voltage Measurement... 10:22:08 Voltage Offset Correction: v_off_ini = 2172.268 uV, v_off_fin = 2194.445 uV 10:22:08 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 227.021 uVs 10:22:08 Integrated Voltage Measurement... 10:22:17 Voltage Offset Correction: v_off_ini = 2189.268 uV, v_off_fin = 2174.671 uV 10:22:17 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -218.970 uVs 10:22:17 Integrated Voltage Measurement... 10:22:26 Voltage Offset Correction: v_off_ini = 2157.054 uV, v_off_fin = 2188.246 uV 10:22:26 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 216.633 uVs 10:22:31 I1Y Average and RMS of Measurements VT = -214.074 +- 3.468 uVs I1Y = -42.812 +- 0.693 uTm 10:22:36 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:22:38 Integrated Voltage Measurement... 10:22:47 Voltage Offset Correction: v_off_ini = 2176.461 uV, v_off_fin = 2175.802 uV 10:22:47 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -77.180 uVs 10:22:47 Integrated Voltage Measurement... 10:22:56 Voltage Offset Correction: v_off_ini = 2182.593 uV, v_off_fin = 2204.175 uV 10:22:56 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 125.821 uVs 10:22:56 Integrated Voltage Measurement... 10:23:05 Voltage Offset Correction: v_off_ini = 2190.970 uV, v_off_fin = 2194.241 uV 10:23:05 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -91.338 uVs 10:23:05 Integrated Voltage Measurement... 10:23:14 Voltage Offset Correction: v_off_ini = 2206.163 uV, v_off_fin = 2223.441 uV 10:23:14 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 81.611 uVs 10:23:14 Integrated Voltage Measurement... 10:23:23 Voltage Offset Correction: v_off_ini = 2212.559 uV, v_off_fin = 2194.890 uV 10:23:23 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -121.557 uVs 10:23:23 Integrated Voltage Measurement... 10:23:32 Voltage Offset Correction: v_off_ini = 2228.828 uV, v_off_fin = 2224.889 uV 10:23:32 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 115.232 uVs 10:23:35 I2X Average and RMS of Measurements VT = -102.123 +- 13.039 uVs I2X = 9.396 +- 1.199 uTm^2 10:23:40 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:23:42 Integrated Voltage Measurement... 10:23:51 Voltage Offset Correction: v_off_ini = 2219.849 uV, v_off_fin = 2202.527 uV 10:23:51 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -206.711 uVs 10:23:51 Integrated Voltage Measurement... 10:24:00 Voltage Offset Correction: v_off_ini = 2209.696 uV, v_off_fin = 2209.029 uV 10:24:00 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 215.697 uVs 10:24:00 Integrated Voltage Measurement... 10:24:09 Voltage Offset Correction: v_off_ini = 2215.183 uV, v_off_fin = 2204.358 uV 10:24:09 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = -208.226 uVs 10:24:10 Integrated Voltage Measurement... 10:24:18 Voltage Offset Correction: v_off_ini = 2200.113 uV, v_off_fin = 2201.055 uV 10:24:18 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 199.398 uVs 10:24:19 Integrated Voltage Measurement... 10:24:28 Voltage Offset Correction: v_off_ini = 2202.108 uV, v_off_fin = 2197.345 uV 10:24:28 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -188.133 uVs 10:24:28 Integrated Voltage Measurement... 10:24:37 Voltage Offset Correction: v_off_ini = 2182.132 uV, v_off_fin = 2209.694 uV 10:24:37 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 236.579 uVs 10:24:39 I2Y Average and RMS of Measurements VT = -209.124 +- 3.785 uVs I2Y = -19.240 +- 0.348 uTm^2 10:24:50 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:24:54 Integrated Voltage Measurement... 10:25:03 Voltage Offset Correction: v_off_ini = 2210.006 uV, v_off_fin = 2211.773 uV 10:25:03 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -113.380 uVs 10:25:03 Integrated Voltage Measurement... 10:25:12 Voltage Offset Correction: v_off_ini = 2195.796 uV, v_off_fin = 2196.182 uV 10:25:12 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 133.707 uVs 10:25:12 Integrated Voltage Measurement... 10:25:21 Voltage Offset Correction: v_off_ini = 2176.620 uV, v_off_fin = 2169.186 uV 10:25:21 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -124.896 uVs 10:25:21 Integrated Voltage Measurement... 10:25:30 Voltage Offset Correction: v_off_ini = 2186.307 uV, v_off_fin = 2176.430 uV 10:25:30 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 111.983 uVs 10:25:30 Integrated Voltage Measurement... 10:25:39 Voltage Offset Correction: v_off_ini = 2197.025 uV, v_off_fin = 2209.157 uV 10:25:39 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -129.415 uVs 10:25:39 Integrated Voltage Measurement... 10:25:48 Voltage Offset Correction: v_off_ini = 2199.363 uV, v_off_fin = 2207.443 uV 10:25:48 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 117.210 uVs 10:25:52 I1X Average and RMS of Measurements VT = -121.765 +- 2.353 uVs I1X = 24.353 +- 0.470 uTm 10:25:55 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:25:59 Integrated Voltage Measurement... 10:26:08 Voltage Offset Correction: v_off_ini = 2191.221 uV, v_off_fin = 2187.692 uV 10:26:08 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -169.488 uVs 10:26:08 Integrated Voltage Measurement... 10:26:17 Voltage Offset Correction: v_off_ini = 2206.250 uV, v_off_fin = 2228.280 uV 10:26:17 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 222.662 uVs 10:26:17 Integrated Voltage Measurement... 10:26:26 Voltage Offset Correction: v_off_ini = 2214.125 uV, v_off_fin = 2208.300 uV 10:26:26 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -181.086 uVs 10:26:26 Integrated Voltage Measurement... 10:26:35 Voltage Offset Correction: v_off_ini = 2192.382 uV, v_off_fin = 2220.331 uV 10:26:35 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 214.962 uVs 10:26:35 Integrated Voltage Measurement... 10:26:44 Voltage Offset Correction: v_off_ini = 2215.788 uV, v_off_fin = 2204.981 uV 10:26:44 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -192.341 uVs 10:26:44 Integrated Voltage Measurement... 10:26:53 Voltage Offset Correction: v_off_ini = 2212.240 uV, v_off_fin = 2224.514 uV 10:26:53 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 199.903 uVs 10:26:57 I1Y Average and RMS of Measurements VT = -196.740 +- 0.908 uVs I1Y = -39.347 +- 0.182 uTm 10:27:02 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:27:05 Integrated Voltage Measurement... 10:27:14 Voltage Offset Correction: v_off_ini = 2198.946 uV, v_off_fin = 2192.404 uV 10:27:14 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -66.926 uVs 10:27:14 Integrated Voltage Measurement... 10:27:23 Voltage Offset Correction: v_off_ini = 2194.999 uV, v_off_fin = 2214.587 uV 10:27:23 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 105.604 uVs 10:27:23 Integrated Voltage Measurement... 10:27:32 Voltage Offset Correction: v_off_ini = 2199.138 uV, v_off_fin = 2199.681 uV 10:27:32 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -122.136 uVs 10:27:32 Integrated Voltage Measurement... 10:27:41 Voltage Offset Correction: v_off_ini = 2184.373 uV, v_off_fin = 2181.808 uV 10:27:41 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 101.815 uVs 10:27:41 Integrated Voltage Measurement... 10:27:50 Voltage Offset Correction: v_off_ini = 2184.015 uV, v_off_fin = 2186.918 uV 10:27:50 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -93.953 uVs 10:27:50 Integrated Voltage Measurement... 10:27:59 Voltage Offset Correction: v_off_ini = 2201.881 uV, v_off_fin = 2212.440 uV 10:27:59 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 87.367 uVs 10:28:01 I2X Average and RMS of Measurements VT = -96.300 +- 11.229 uVs I2X = 8.860 +- 1.033 uTm^2 10:28:06 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:28:09 Integrated Voltage Measurement... 10:28:18 Voltage Offset Correction: v_off_ini = 2206.027 uV, v_off_fin = 2204.207 uV 10:28:18 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -216.495 uVs 10:28:18 Integrated Voltage Measurement... 10:28:27 Voltage Offset Correction: v_off_ini = 2215.868 uV, v_off_fin = 2209.033 uV 10:28:27 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 205.792 uVs 10:28:27 Integrated Voltage Measurement... 10:28:36 Voltage Offset Correction: v_off_ini = 2200.994 uV, v_off_fin = 2196.620 uV 10:28:36 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -205.883 uVs 10:28:36 Integrated Voltage Measurement... 10:28:45 Voltage Offset Correction: v_off_ini = 2212.110 uV, v_off_fin = 2202.662 uV 10:28:45 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 205.328 uVs 10:28:45 Integrated Voltage Measurement... 10:28:54 Voltage Offset Correction: v_off_ini = 2190.685 uV, v_off_fin = 2195.625 uV 10:28:54 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -220.369 uVs 10:28:54 Integrated Voltage Measurement... 10:29:03 Voltage Offset Correction: v_off_ini = 2194.197 uV, v_off_fin = 2191.204 uV 10:29:03 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = 215.552 uVs 10:29:05 I2Y Average and RMS of Measurements VT = -211.570 +- 5.053 uVs I2Y = -19.465 +- 0.465 uTm^2 10:29:16 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:29:20 Integrated Voltage Measurement... 10:29:29 Voltage Offset Correction: v_off_ini = 2205.402 uV, v_off_fin = 2200.753 uV 10:29:29 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -118.512 uVs 10:29:29 Integrated Voltage Measurement... 10:29:38 Voltage Offset Correction: v_off_ini = 2209.698 uV, v_off_fin = 2209.675 uV 10:29:38 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 139.819 uVs 10:29:38 Integrated Voltage Measurement... 10:29:47 Voltage Offset Correction: v_off_ini = 2216.212 uV, v_off_fin = 2238.542 uV 10:29:47 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -116.573 uVs 10:29:47 Integrated Voltage Measurement... 10:29:56 Voltage Offset Correction: v_off_ini = 2221.041 uV, v_off_fin = 2207.882 uV 10:29:56 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 118.139 uVs 10:29:56 Integrated Voltage Measurement... 10:30:05 Voltage Offset Correction: v_off_ini = 2203.763 uV, v_off_fin = 2201.795 uV 10:30:05 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -136.568 uVs 10:30:05 Integrated Voltage Measurement... 10:30:14 Voltage Offset Correction: v_off_ini = 2199.550 uV, v_off_fin = 2218.575 uV 10:30:14 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 122.880 uVs 10:30:18 I1X Average and RMS of Measurements VT = -125.415 +- 5.703 uVs I1X = 25.083 +- 1.141 uTm 10:30:21 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:30:25 Integrated Voltage Measurement... 10:30:34 Voltage Offset Correction: v_off_ini = 2187.097 uV, v_off_fin = 2172.267 uV 10:30:34 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -197.818 uVs 10:30:34 Integrated Voltage Measurement... 10:30:43 Voltage Offset Correction: v_off_ini = 2201.348 uV, v_off_fin = 2209.328 uV 10:30:43 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 211.823 uVs 10:30:43 Integrated Voltage Measurement... 10:30:52 Voltage Offset Correction: v_off_ini = 2186.884 uV, v_off_fin = 2189.077 uV 10:30:52 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -177.366 uVs 10:30:52 Integrated Voltage Measurement... 10:31:01 Voltage Offset Correction: v_off_ini = 2191.126 uV, v_off_fin = 2212.841 uV 10:31:01 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 203.110 uVs 10:31:01 Integrated Voltage Measurement... 10:31:10 Voltage Offset Correction: v_off_ini = 2192.245 uV, v_off_fin = 2188.069 uV 10:31:10 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -175.126 uVs 10:31:10 Integrated Voltage Measurement... 10:31:19 Voltage Offset Correction: v_off_ini = 2204.030 uV, v_off_fin = 2204.564 uV 10:31:19 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 209.826 uVs 10:31:23 I1Y Average and RMS of Measurements VT = -195.845 +- 6.412 uVs I1Y = -39.166 +- 1.283 uTm 10:31:28 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:31:31 Integrated Voltage Measurement... 10:31:40 Voltage Offset Correction: v_off_ini = 2199.057 uV, v_off_fin = 2202.565 uV 10:31:40 Integrated Voltage Measurement dXYus = -1.0001 mm, dXYds = 0.0000 mm dXY = -1.0001 mm VT = -93.379 uVs 10:31:40 Integrated Voltage Measurement... 10:31:49 Voltage Offset Correction: v_off_ini = 2202.752 uV, v_off_fin = 2220.547 uV 10:31:49 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 83.693 uVs 10:31:49 Integrated Voltage Measurement... 10:31:58 Voltage Offset Correction: v_off_ini = 2200.783 uV, v_off_fin = 2198.335 uV 10:31:58 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -109.245 uVs 10:31:58 Integrated Voltage Measurement... 10:32:07 Voltage Offset Correction: v_off_ini = 2204.375 uV, v_off_fin = 2191.477 uV 10:32:07 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 135.902 uVs 10:32:07 Integrated Voltage Measurement... 10:32:16 Voltage Offset Correction: v_off_ini = 2198.481 uV, v_off_fin = 2196.596 uV 10:32:16 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -96.552 uVs 10:32:16 Integrated Voltage Measurement... 10:32:25 Voltage Offset Correction: v_off_ini = 2205.321 uV, v_off_fin = 2227.008 uV 10:32:25 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 121.952 uVs 10:32:28 I2X Average and RMS of Measurements VT = -106.787 +- 14.005 uVs I2X = 9.824 +- 1.289 uTm^2 10:32:33 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:32:35 Integrated Voltage Measurement... 10:32:44 Voltage Offset Correction: v_off_ini = 2215.166 uV, v_off_fin = 2223.435 uV 10:32:44 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -209.839 uVs 10:32:44 Integrated Voltage Measurement... 10:32:53 Voltage Offset Correction: v_off_ini = 2217.615 uV, v_off_fin = 2223.337 uV 10:32:53 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 206.285 uVs 10:32:53 Integrated Voltage Measurement... 10:33:02 Voltage Offset Correction: v_off_ini = 2256.625 uV, v_off_fin = 2243.753 uV 10:33:02 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -225.747 uVs 10:33:02 Integrated Voltage Measurement... 10:33:11 Voltage Offset Correction: v_off_ini = 2237.377 uV, v_off_fin = 2223.053 uV 10:33:11 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 227.257 uVs 10:33:11 Integrated Voltage Measurement... 10:33:20 Voltage Offset Correction: v_off_ini = 2230.556 uV, v_off_fin = 2228.405 uV 10:33:20 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -213.736 uVs 10:33:20 Integrated Voltage Measurement... 10:33:29 Voltage Offset Correction: v_off_ini = 2237.748 uV, v_off_fin = 2237.354 uV 10:33:29 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 190.700 uVs 10:33:32 I2Y Average and RMS of Measurements VT = -212.261 +- 10.349 uVs I2Y = -19.528 +- 0.952 uTm^2 10:33:42 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:33:47 Integrated Voltage Measurement... 10:33:56 Voltage Offset Correction: v_off_ini = 2251.658 uV, v_off_fin = 2253.320 uV 10:33:56 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -152.720 uVs 10:33:56 Integrated Voltage Measurement... 10:34:05 Voltage Offset Correction: v_off_ini = 2275.036 uV, v_off_fin = 2272.752 uV 10:34:05 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 107.514 uVs 10:34:05 Integrated Voltage Measurement... 10:34:14 Voltage Offset Correction: v_off_ini = 2237.324 uV, v_off_fin = 2252.766 uV 10:34:14 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -86.532 uVs 10:34:14 Integrated Voltage Measurement... 10:34:23 Voltage Offset Correction: v_off_ini = 2264.780 uV, v_off_fin = 2250.540 uV 10:34:23 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 79.538 uVs 10:34:23 Integrated Voltage Measurement... 10:34:32 Voltage Offset Correction: v_off_ini = 2253.996 uV, v_off_fin = 2247.839 uV 10:34:32 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -124.570 uVs 10:34:32 Integrated Voltage Measurement... 10:34:41 Voltage Offset Correction: v_off_ini = 2257.468 uV, v_off_fin = 2271.892 uV 10:34:41 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 117.329 uVs 10:34:45 I1X Average and RMS of Measurements VT = -111.367 +- 20.380 uVs I1X = 22.273 +- 4.076 uTm 10:34:47 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:34:52 Integrated Voltage Measurement... 10:35:01 Voltage Offset Correction: v_off_ini = 2272.888 uV, v_off_fin = 2267.251 uV 10:35:01 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -185.235 uVs 10:35:01 Integrated Voltage Measurement... 10:35:10 Voltage Offset Correction: v_off_ini = 2289.733 uV, v_off_fin = 2283.899 uV 10:35:10 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 181.460 uVs 10:35:10 Integrated Voltage Measurement... 10:35:19 Voltage Offset Correction: v_off_ini = 2295.081 uV, v_off_fin = 2282.898 uV 10:35:19 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -207.441 uVs 10:35:19 Integrated Voltage Measurement... 10:35:28 Voltage Offset Correction: v_off_ini = 2287.852 uV, v_off_fin = 2285.402 uV 10:35:28 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 167.069 uVs 10:35:28 Integrated Voltage Measurement... 10:35:37 Voltage Offset Correction: v_off_ini = 2289.453 uV, v_off_fin = 2281.442 uV 10:35:37 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -185.296 uVs 10:35:37 Integrated Voltage Measurement... 10:35:46 Voltage Offset Correction: v_off_ini = 2273.695 uV, v_off_fin = 2280.188 uV 10:35:46 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 197.586 uVs 10:35:50 I1Y Average and RMS of Measurements VT = -187.348 +- 3.305 uVs I1Y = -37.468 +- 0.661 uTm 10:35:55 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:35:57 Integrated Voltage Measurement... 10:36:06 Voltage Offset Correction: v_off_ini = 2263.730 uV, v_off_fin = 2268.253 uV 10:36:06 Integrated Voltage Measurement dXYus = -1.0001 mm, dXYds = 0.0000 mm dXY = -1.0001 mm VT = -106.327 uVs 10:36:06 Integrated Voltage Measurement... 10:36:15 Voltage Offset Correction: v_off_ini = 2295.595 uV, v_off_fin = 2290.781 uV 10:36:15 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 73.419 uVs 10:36:15 Integrated Voltage Measurement... 10:36:24 Voltage Offset Correction: v_off_ini = 2288.878 uV, v_off_fin = 2281.932 uV 10:36:24 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -105.093 uVs 10:36:24 Integrated Voltage Measurement... 10:36:33 Voltage Offset Correction: v_off_ini = 2283.808 uV, v_off_fin = 2293.255 uV 10:36:33 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 131.964 uVs 10:36:34 Integrated Voltage Measurement... 10:36:42 Voltage Offset Correction: v_off_ini = 2288.333 uV, v_off_fin = 2286.444 uV 10:36:43 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -108.788 uVs 10:36:43 Integrated Voltage Measurement... 10:36:52 Voltage Offset Correction: v_off_ini = 2277.476 uV, v_off_fin = 2288.269 uV 10:36:52 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 65.950 uVs 10:36:54 I2X Average and RMS of Measurements VT = -98.590 +- 14.136 uVs I2X = 9.070 +- 1.301 uTm^2 10:36:59 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:37:02 Integrated Voltage Measurement... 10:37:11 Voltage Offset Correction: v_off_ini = 2291.764 uV, v_off_fin = 2283.732 uV 10:37:11 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -202.342 uVs 10:37:11 Integrated Voltage Measurement... 10:37:20 Voltage Offset Correction: v_off_ini = 2286.281 uV, v_off_fin = 2288.074 uV 10:37:20 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 210.921 uVs 10:37:20 Integrated Voltage Measurement... 10:37:29 Voltage Offset Correction: v_off_ini = 2291.789 uV, v_off_fin = 2291.738 uV 10:37:29 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0001 mm dXY = 1.0000 mm VT = -193.189 uVs 10:37:29 Integrated Voltage Measurement... 10:37:38 Voltage Offset Correction: v_off_ini = 2302.031 uV, v_off_fin = 2307.900 uV 10:37:38 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = -0.0001 mm dXY = -1.0000 mm VT = 204.262 uVs 10:37:38 Integrated Voltage Measurement... 10:37:47 Voltage Offset Correction: v_off_ini = 2328.761 uV, v_off_fin = 2328.410 uV 10:37:47 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -189.768 uVs 10:37:47 Integrated Voltage Measurement... 10:37:56 Voltage Offset Correction: v_off_ini = 2323.094 uV, v_off_fin = 2319.975 uV 10:37:56 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 205.158 uVs 10:37:58 I2Y Average and RMS of Measurements VT = -200.940 +- 4.057 uVs I2Y = -18.486 +- 0.373 uTm^2 10:38:09 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:38:13 Integrated Voltage Measurement... 10:38:22 Voltage Offset Correction: v_off_ini = 2354.666 uV, v_off_fin = 2336.813 uV 10:38:22 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -118.689 uVs 10:38:22 Integrated Voltage Measurement... 10:38:31 Voltage Offset Correction: v_off_ini = 2338.175 uV, v_off_fin = 2335.557 uV 10:38:31 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 114.669 uVs 10:38:31 Integrated Voltage Measurement... 10:38:40 Voltage Offset Correction: v_off_ini = 2324.438 uV, v_off_fin = 2329.742 uV 10:38:40 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -120.404 uVs 10:38:40 Integrated Voltage Measurement... 10:38:49 Voltage Offset Correction: v_off_ini = 2342.506 uV, v_off_fin = 2347.497 uV 10:38:49 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 115.873 uVs 10:38:49 Integrated Voltage Measurement... 10:38:58 Voltage Offset Correction: v_off_ini = 2343.366 uV, v_off_fin = 2345.294 uV 10:38:58 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -122.617 uVs 10:38:58 Integrated Voltage Measurement... 10:39:07 Voltage Offset Correction: v_off_ini = 2360.841 uV, v_off_fin = 2379.376 uV 10:39:07 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 133.566 uVs 10:39:11 I1X Average and RMS of Measurements VT = -120.969 +- 5.071 uVs I1X = 24.194 +- 1.014 uTm 10:39:14 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:39:18 Integrated Voltage Measurement... 10:39:27 Voltage Offset Correction: v_off_ini = 2330.677 uV, v_off_fin = 2332.443 uV 10:39:27 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -183.181 uVs 10:39:27 Integrated Voltage Measurement... 10:39:36 Voltage Offset Correction: v_off_ini = 2340.273 uV, v_off_fin = 2356.901 uV 10:39:36 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5002 mm dXY = -0.5001 mm VT = 196.584 uVs 10:39:36 Integrated Voltage Measurement... 10:39:45 Voltage Offset Correction: v_off_ini = 2344.322 uV, v_off_fin = 2343.367 uV 10:39:45 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -186.201 uVs 10:39:45 Integrated Voltage Measurement... 10:39:54 Voltage Offset Correction: v_off_ini = 2351.192 uV, v_off_fin = 2341.015 uV 10:39:54 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 187.445 uVs 10:39:54 Integrated Voltage Measurement... 10:40:03 Voltage Offset Correction: v_off_ini = 2343.591 uV, v_off_fin = 2335.996 uV 10:40:03 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -172.900 uVs 10:40:03 Integrated Voltage Measurement... 10:40:12 Voltage Offset Correction: v_off_ini = 2338.471 uV, v_off_fin = 2354.426 uV 10:40:12 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5002 mm dXY = -0.5001 mm VT = 183.731 uVs 10:40:16 I1Y Average and RMS of Measurements VT = -185.007 +- 4.894 uVs I1Y = -36.998 +- 0.979 uTm 10:40:21 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:40:24 Integrated Voltage Measurement... 10:40:33 Voltage Offset Correction: v_off_ini = 2326.582 uV, v_off_fin = 2319.250 uV 10:40:33 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -68.553 uVs 10:40:33 Integrated Voltage Measurement... 10:40:42 Voltage Offset Correction: v_off_ini = 2331.757 uV, v_off_fin = 2362.442 uV 10:40:42 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 90.114 uVs 10:40:42 Integrated Voltage Measurement... 10:40:51 Voltage Offset Correction: v_off_ini = 2340.539 uV, v_off_fin = 2326.504 uV 10:40:51 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -103.081 uVs 10:40:51 Integrated Voltage Measurement... 10:41:00 Voltage Offset Correction: v_off_ini = 2338.111 uV, v_off_fin = 2359.558 uV 10:41:00 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 83.772 uVs 10:41:00 Integrated Voltage Measurement... 10:41:09 Voltage Offset Correction: v_off_ini = 2350.260 uV, v_off_fin = 2326.637 uV 10:41:09 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -89.542 uVs 10:41:09 Integrated Voltage Measurement... 10:41:18 Voltage Offset Correction: v_off_ini = 2324.555 uV, v_off_fin = 2359.591 uV 10:41:18 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 123.602 uVs 10:41:20 I2X Average and RMS of Measurements VT = -93.111 +- 11.123 uVs I2X = 8.566 +- 1.023 uTm^2 10:41:25 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:41:28 Integrated Voltage Measurement... 10:41:37 Voltage Offset Correction: v_off_ini = 2322.729 uV, v_off_fin = 2334.523 uV 10:41:37 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -176.324 uVs 10:41:37 Integrated Voltage Measurement... 10:41:46 Voltage Offset Correction: v_off_ini = 2314.283 uV, v_off_fin = 2329.615 uV 10:41:46 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 186.256 uVs 10:41:46 Integrated Voltage Measurement... 10:41:55 Voltage Offset Correction: v_off_ini = 2309.089 uV, v_off_fin = 2288.046 uV 10:41:55 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -182.407 uVs 10:41:55 Integrated Voltage Measurement... 10:42:04 Voltage Offset Correction: v_off_ini = 2307.588 uV, v_off_fin = 2324.478 uV 10:42:04 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 190.144 uVs 10:42:04 Integrated Voltage Measurement... 10:42:13 Voltage Offset Correction: v_off_ini = 2319.642 uV, v_off_fin = 2317.822 uV 10:42:13 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -192.053 uVs 10:42:13 Integrated Voltage Measurement... 10:42:22 Voltage Offset Correction: v_off_ini = 2326.872 uV, v_off_fin = 2336.264 uV 10:42:22 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 196.717 uVs 10:42:25 I2Y Average and RMS of Measurements VT = -187.317 +- 5.396 uVs I2Y = -17.233 +- 0.496 uTm^2 10:42:35 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:42:40 Integrated Voltage Measurement... 10:42:49 Voltage Offset Correction: v_off_ini = 2333.063 uV, v_off_fin = 2329.370 uV 10:42:49 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -112.801 uVs 10:42:49 Integrated Voltage Measurement... 10:42:58 Voltage Offset Correction: v_off_ini = 2364.939 uV, v_off_fin = 2357.217 uV 10:42:58 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 106.415 uVs 10:42:58 Integrated Voltage Measurement... 10:43:07 Voltage Offset Correction: v_off_ini = 2356.146 uV, v_off_fin = 2335.510 uV 10:43:07 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -115.075 uVs 10:43:07 Integrated Voltage Measurement... 10:43:16 Voltage Offset Correction: v_off_ini = 2347.511 uV, v_off_fin = 2358.253 uV 10:43:16 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 132.670 uVs 10:43:16 Integrated Voltage Measurement... 10:43:25 Voltage Offset Correction: v_off_ini = 2344.234 uV, v_off_fin = 2352.789 uV 10:43:25 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -121.222 uVs 10:43:25 Integrated Voltage Measurement... 10:43:34 Voltage Offset Correction: v_off_ini = 2341.882 uV, v_off_fin = 2357.347 uV 10:43:34 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 113.033 uVs 10:43:38 I1X Average and RMS of Measurements VT = -116.869 +- 5.826 uVs I1X = 23.374 +- 1.165 uTm 10:43:40 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:43:45 Integrated Voltage Measurement... 10:43:54 Voltage Offset Correction: v_off_ini = 2350.177 uV, v_off_fin = 2343.782 uV 10:43:54 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -174.921 uVs 10:43:54 Integrated Voltage Measurement... 10:44:03 Voltage Offset Correction: v_off_ini = 2349.439 uV, v_off_fin = 2381.694 uV 10:44:03 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5002 mm dXY = -0.5001 mm VT = 191.238 uVs 10:44:03 Integrated Voltage Measurement... 10:44:12 Voltage Offset Correction: v_off_ini = 2364.673 uV, v_off_fin = 2367.071 uV 10:44:12 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5002 mm dXY = 0.5001 mm VT = -171.932 uVs 10:44:12 Integrated Voltage Measurement... 10:44:20 Voltage Offset Correction: v_off_ini = 2370.506 uV, v_off_fin = 2362.081 uV 10:44:20 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.5002 mm dXY = -0.5000 mm VT = 170.240 uVs 10:44:21 Integrated Voltage Measurement... 10:44:29 Voltage Offset Correction: v_off_ini = 2353.299 uV, v_off_fin = 2342.457 uV 10:44:29 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -168.157 uVs 10:44:30 Integrated Voltage Measurement... 10:44:38 Voltage Offset Correction: v_off_ini = 2383.231 uV, v_off_fin = 2377.339 uV 10:44:38 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.5002 mm dXY = -0.5000 mm VT = 164.081 uVs 10:44:43 I1Y Average and RMS of Measurements VT = -173.428 +- 7.120 uVs I1Y = -34.682 +- 1.424 uTm 10:44:48 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:44:50 Integrated Voltage Measurement... 10:44:59 Voltage Offset Correction: v_off_ini = 2388.489 uV, v_off_fin = 2369.578 uV 10:44:59 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -69.146 uVs 10:44:59 Integrated Voltage Measurement... 10:45:08 Voltage Offset Correction: v_off_ini = 2383.028 uV, v_off_fin = 2396.769 uV 10:45:08 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 114.115 uVs 10:45:08 Integrated Voltage Measurement... 10:45:17 Voltage Offset Correction: v_off_ini = 2382.178 uV, v_off_fin = 2381.005 uV 10:45:17 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -81.744 uVs 10:45:17 Integrated Voltage Measurement... 10:45:26 Voltage Offset Correction: v_off_ini = 2361.003 uV, v_off_fin = 2394.975 uV 10:45:26 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 89.158 uVs 10:45:26 Integrated Voltage Measurement... 10:45:35 Voltage Offset Correction: v_off_ini = 2398.079 uV, v_off_fin = 2390.475 uV 10:45:35 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -80.019 uVs 10:45:35 Integrated Voltage Measurement... 10:45:44 Voltage Offset Correction: v_off_ini = 2384.984 uV, v_off_fin = 2375.885 uV 10:45:44 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 97.511 uVs 10:45:47 I2X Average and RMS of Measurements VT = -88.616 +- 2.525 uVs I2X = 8.153 +- 0.232 uTm^2 10:45:52 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:45:55 Integrated Voltage Measurement... 10:46:03 Voltage Offset Correction: v_off_ini = 2419.075 uV, v_off_fin = 2409.791 uV 10:46:04 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -205.472 uVs 10:46:04 Integrated Voltage Measurement... 10:46:13 Voltage Offset Correction: v_off_ini = 2408.337 uV, v_off_fin = 2414.379 uV 10:46:13 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 171.356 uVs 10:46:13 Integrated Voltage Measurement... 10:46:22 Voltage Offset Correction: v_off_ini = 2402.677 uV, v_off_fin = 2395.849 uV 10:46:22 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -183.916 uVs 10:46:22 Integrated Voltage Measurement... 10:46:31 Voltage Offset Correction: v_off_ini = 2401.872 uV, v_off_fin = 2421.718 uV 10:46:31 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 200.112 uVs 10:46:31 Integrated Voltage Measurement... 10:46:40 Voltage Offset Correction: v_off_ini = 2454.455 uV, v_off_fin = 2436.557 uV 10:46:40 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -192.392 uVs 10:46:40 Integrated Voltage Measurement... 10:46:49 Voltage Offset Correction: v_off_ini = 2433.541 uV, v_off_fin = 2432.153 uV 10:46:49 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 170.990 uVs 10:46:51 I2Y Average and RMS of Measurements VT = -187.373 +- 4.278 uVs I2Y = -17.238 +- 0.394 uTm^2 10:47:02 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:47:06 Integrated Voltage Measurement... 10:47:15 Voltage Offset Correction: v_off_ini = 2420.161 uV, v_off_fin = 2414.242 uV 10:47:15 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -122.288 uVs 10:47:15 Integrated Voltage Measurement... 10:47:24 Voltage Offset Correction: v_off_ini = 2395.657 uV, v_off_fin = 2399.267 uV 10:47:24 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 113.486 uVs 10:47:24 Integrated Voltage Measurement... 10:47:33 Voltage Offset Correction: v_off_ini = 2399.576 uV, v_off_fin = 2397.393 uV 10:47:33 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -115.897 uVs 10:47:33 Integrated Voltage Measurement... 10:47:42 Voltage Offset Correction: v_off_ini = 2407.427 uV, v_off_fin = 2421.191 uV 10:47:42 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 82.621 uVs 10:47:42 Integrated Voltage Measurement... 10:47:51 Voltage Offset Correction: v_off_ini = 2407.484 uV, v_off_fin = 2410.771 uV 10:47:51 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -106.370 uVs 10:47:51 Integrated Voltage Measurement... 10:48:00 Voltage Offset Correction: v_off_ini = 2395.072 uV, v_off_fin = 2408.450 uV 10:48:00 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 117.726 uVs 10:48:04 I1X Average and RMS of Measurements VT = -109.731 +- 7.779 uVs I1X = 21.947 +- 1.556 uTm 10:48:07 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:48:11 Integrated Voltage Measurement... 10:48:20 Voltage Offset Correction: v_off_ini = 2419.366 uV, v_off_fin = 2431.713 uV 10:48:20 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -172.209 uVs 10:48:20 Integrated Voltage Measurement... 10:48:29 Voltage Offset Correction: v_off_ini = 2421.986 uV, v_off_fin = 2414.306 uV 10:48:29 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5002 mm dXY = -0.5001 mm VT = 157.509 uVs 10:48:29 Integrated Voltage Measurement... 10:48:38 Voltage Offset Correction: v_off_ini = 2421.247 uV, v_off_fin = 2405.291 uV 10:48:38 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -170.346 uVs 10:48:38 Integrated Voltage Measurement... 10:48:47 Voltage Offset Correction: v_off_ini = 2410.778 uV, v_off_fin = 2423.521 uV 10:48:47 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 177.774 uVs 10:48:47 Integrated Voltage Measurement... 10:48:56 Voltage Offset Correction: v_off_ini = 2405.909 uV, v_off_fin = 2393.792 uV 10:48:56 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -164.420 uVs 10:48:56 Integrated Voltage Measurement... 10:49:05 Voltage Offset Correction: v_off_ini = 2413.831 uV, v_off_fin = 2410.314 uV 10:49:05 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 168.916 uVs 10:49:09 I1Y Average and RMS of Measurements VT = -168.529 +- 3.980 uVs I1Y = -33.702 +- 0.796 uTm 10:49:14 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:49:17 Integrated Voltage Measurement... 10:49:26 Voltage Offset Correction: v_off_ini = 2427.466 uV, v_off_fin = 2426.130 uV 10:49:26 Integrated Voltage Measurement dXYus = -0.9998 mm, dXYds = 0.0000 mm dXY = -0.9998 mm VT = -80.543 uVs 10:49:26 Integrated Voltage Measurement... 10:49:35 Voltage Offset Correction: v_off_ini = 2433.588 uV, v_off_fin = 2440.076 uV 10:49:35 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 117.002 uVs 10:49:35 Integrated Voltage Measurement... 10:49:44 Voltage Offset Correction: v_off_ini = 2442.046 uV, v_off_fin = 2438.743 uV 10:49:44 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -95.381 uVs 10:49:44 Integrated Voltage Measurement... 10:49:53 Voltage Offset Correction: v_off_ini = 2447.707 uV, v_off_fin = 2452.625 uV 10:49:53 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 67.810 uVs 10:49:53 Integrated Voltage Measurement... 10:50:02 Voltage Offset Correction: v_off_ini = 2434.679 uV, v_off_fin = 2416.728 uV 10:50:02 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -98.480 uVs 10:50:02 Integrated Voltage Measurement... 10:50:11 Voltage Offset Correction: v_off_ini = 2435.405 uV, v_off_fin = 2430.609 uV 10:50:11 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 83.952 uVs 10:50:13 I2X Average and RMS of Measurements VT = -90.528 +- 7.029 uVs I2X = 8.329 +- 0.647 uTm^2 10:50:18 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:50:21 Integrated Voltage Measurement... 10:50:30 Voltage Offset Correction: v_off_ini = 2444.068 uV, v_off_fin = 2428.413 uV 10:50:30 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -188.421 uVs 10:50:30 Integrated Voltage Measurement... 10:50:39 Voltage Offset Correction: v_off_ini = 2402.851 uV, v_off_fin = 2403.614 uV 10:50:39 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 180.629 uVs 10:50:39 Integrated Voltage Measurement... 10:50:48 Voltage Offset Correction: v_off_ini = 2425.679 uV, v_off_fin = 2433.619 uV 10:50:48 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -159.366 uVs 10:50:48 Integrated Voltage Measurement... 10:50:57 Voltage Offset Correction: v_off_ini = 2444.003 uV, v_off_fin = 2450.454 uV 10:50:57 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 163.222 uVs 10:50:57 Integrated Voltage Measurement... 10:51:06 Voltage Offset Correction: v_off_ini = 2447.001 uV, v_off_fin = 2441.976 uV 10:51:06 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -182.572 uVs 10:51:06 Integrated Voltage Measurement... 10:51:15 Voltage Offset Correction: v_off_ini = 2439.146 uV, v_off_fin = 2435.250 uV 10:51:15 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 160.188 uVs 10:51:18 I2Y Average and RMS of Measurements VT = -172.400 +- 9.511 uVs I2Y = -15.861 +- 0.875 uTm^2 10:51:28 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:51:33 Integrated Voltage Measurement... 10:51:42 Voltage Offset Correction: v_off_ini = 2444.803 uV, v_off_fin = 2430.222 uV 10:51:42 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -128.589 uVs 10:51:42 Integrated Voltage Measurement... 10:51:51 Voltage Offset Correction: v_off_ini = 2421.641 uV, v_off_fin = 2428.628 uV 10:51:51 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 91.939 uVs 10:51:51 Integrated Voltage Measurement... 10:52:00 Voltage Offset Correction: v_off_ini = 2434.682 uV, v_off_fin = 2436.488 uV 10:52:00 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.4999 mm dXY = -0.4999 mm VT = -93.833 uVs 10:52:00 Integrated Voltage Measurement... 10:52:09 Voltage Offset Correction: v_off_ini = 2441.752 uV, v_off_fin = 2443.470 uV 10:52:09 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = 96.997 uVs 10:52:09 Integrated Voltage Measurement... 10:52:18 Voltage Offset Correction: v_off_ini = 2436.155 uV, v_off_fin = 2435.354 uV 10:52:18 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -105.010 uVs 10:52:18 Integrated Voltage Measurement... 10:52:27 Voltage Offset Correction: v_off_ini = 2427.068 uV, v_off_fin = 2428.729 uV 10:52:27 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 95.482 uVs 10:52:31 I1X Average and RMS of Measurements VT = -101.975 +- 6.184 uVs I1X = 20.396 +- 1.236 uTm 10:52:33 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:52:38 Integrated Voltage Measurement... 10:52:47 Voltage Offset Correction: v_off_ini = 2423.921 uV, v_off_fin = 2411.481 uV 10:52:47 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = -160.262 uVs 10:52:47 Integrated Voltage Measurement... 10:52:56 Voltage Offset Correction: v_off_ini = 2431.436 uV, v_off_fin = 2456.696 uV 10:52:56 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 165.752 uVs 10:52:56 Integrated Voltage Measurement... 10:53:05 Voltage Offset Correction: v_off_ini = 2424.689 uV, v_off_fin = 2409.033 uV 10:53:05 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5002 mm dXY = 0.5001 mm VT = -163.437 uVs 10:53:05 Integrated Voltage Measurement... 10:53:14 Voltage Offset Correction: v_off_ini = 2410.446 uV, v_off_fin = 2431.424 uV 10:53:14 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5002 mm dXY = -0.5001 mm VT = 162.151 uVs 10:53:14 Integrated Voltage Measurement... 10:53:22 Voltage Offset Correction: v_off_ini = 2420.553 uV, v_off_fin = 2424.673 uV 10:53:23 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5002 mm dXY = 0.5001 mm VT = -175.735 uVs 10:53:23 Integrated Voltage Measurement... 10:53:31 Voltage Offset Correction: v_off_ini = 2422.256 uV, v_off_fin = 2438.807 uV 10:53:31 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5002 mm dXY = -0.5001 mm VT = 181.628 uVs 10:53:36 I1Y Average and RMS of Measurements VT = -168.161 +- 7.440 uVs I1Y = -33.628 +- 1.486 uTm 10:53:41 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:53:43 Integrated Voltage Measurement... 10:53:52 Voltage Offset Correction: v_off_ini = 2471.934 uV, v_off_fin = 2460.630 uV 10:53:52 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -89.596 uVs 10:53:52 Integrated Voltage Measurement... 10:54:01 Voltage Offset Correction: v_off_ini = 2447.198 uV, v_off_fin = 2448.167 uV 10:54:01 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 99.732 uVs 10:54:01 Integrated Voltage Measurement... 10:54:10 Voltage Offset Correction: v_off_ini = 2457.421 uV, v_off_fin = 2459.913 uV 10:54:10 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -101.773 uVs 10:54:10 Integrated Voltage Measurement... 10:54:19 Voltage Offset Correction: v_off_ini = 2455.855 uV, v_off_fin = 2474.263 uV 10:54:19 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 103.473 uVs 10:54:20 Integrated Voltage Measurement... 10:54:28 Voltage Offset Correction: v_off_ini = 2443.825 uV, v_off_fin = 2430.549 uV 10:54:29 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -102.627 uVs 10:54:29 Integrated Voltage Measurement... 10:54:38 Voltage Offset Correction: v_off_ini = 2434.687 uV, v_off_fin = 2434.548 uV 10:54:38 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 98.358 uVs 10:54:40 I2X Average and RMS of Measurements VT = -99.260 +- 3.364 uVs I2X = 9.132 +- 0.309 uTm^2 10:54:45 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:54:48 Integrated Voltage Measurement... 10:54:57 Voltage Offset Correction: v_off_ini = 2428.477 uV, v_off_fin = 2430.239 uV 10:54:57 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -171.788 uVs 10:54:57 Integrated Voltage Measurement... 10:55:06 Voltage Offset Correction: v_off_ini = 2424.930 uV, v_off_fin = 2420.253 uV 10:55:06 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 146.317 uVs 10:55:06 Integrated Voltage Measurement... 10:55:15 Voltage Offset Correction: v_off_ini = 2433.409 uV, v_off_fin = 2433.295 uV 10:55:15 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -156.414 uVs 10:55:15 Integrated Voltage Measurement... 10:55:24 Voltage Offset Correction: v_off_ini = 2422.414 uV, v_off_fin = 2447.431 uV 10:55:24 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 165.902 uVs 10:55:24 Integrated Voltage Measurement... 10:55:33 Voltage Offset Correction: v_off_ini = 2442.726 uV, v_off_fin = 2452.528 uV 10:55:33 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -179.445 uVs 10:55:33 Integrated Voltage Measurement... 10:55:42 Voltage Offset Correction: v_off_ini = 2461.891 uV, v_off_fin = 2458.560 uV 10:55:42 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 174.342 uVs 10:55:44 I2Y Average and RMS of Measurements VT = -165.701 +- 7.961 uVs I2Y = -15.245 +- 0.732 uTm^2 10:55:55 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:55:59 Integrated Voltage Measurement... 10:56:08 Voltage Offset Correction: v_off_ini = 2438.505 uV, v_off_fin = 2444.871 uV 10:56:08 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.4998 mm dXY = -0.4999 mm VT = -113.569 uVs 10:56:08 Integrated Voltage Measurement... 10:56:17 Voltage Offset Correction: v_off_ini = 2445.296 uV, v_off_fin = 2472.660 uV 10:56:17 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 92.578 uVs 10:56:17 Integrated Voltage Measurement... 10:56:26 Voltage Offset Correction: v_off_ini = 2485.510 uV, v_off_fin = 2472.889 uV 10:56:26 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -110.049 uVs 10:56:26 Integrated Voltage Measurement... 10:56:35 Voltage Offset Correction: v_off_ini = 2486.753 uV, v_off_fin = 2470.790 uV 10:56:35 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = 85.297 uVs 10:56:35 Integrated Voltage Measurement... 10:56:44 Voltage Offset Correction: v_off_ini = 2468.538 uV, v_off_fin = 2437.494 uV 10:56:44 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -123.452 uVs 10:56:44 Integrated Voltage Measurement... 10:56:53 Voltage Offset Correction: v_off_ini = 2472.617 uV, v_off_fin = 2497.709 uV 10:56:53 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 106.328 uVs 10:56:57 I1X Average and RMS of Measurements VT = -105.212 +- 7.190 uVs I1X = 21.045 +- 1.437 uTm 10:57:00 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:57:04 Integrated Voltage Measurement... 10:57:13 Voltage Offset Correction: v_off_ini = 2466.640 uV, v_off_fin = 2483.073 uV 10:57:13 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -121.250 uVs 10:57:13 Integrated Voltage Measurement... 10:57:22 Voltage Offset Correction: v_off_ini = 2489.526 uV, v_off_fin = 2491.345 uV 10:57:22 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5002 mm dXY = -0.5001 mm VT = 158.022 uVs 10:57:22 Integrated Voltage Measurement... 10:57:31 Voltage Offset Correction: v_off_ini = 2475.548 uV, v_off_fin = 2464.114 uV 10:57:31 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5002 mm dXY = 0.5001 mm VT = -147.293 uVs 10:57:31 Integrated Voltage Measurement... 10:57:40 Voltage Offset Correction: v_off_ini = 2476.640 uV, v_off_fin = 2488.806 uV 10:57:40 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 166.063 uVs 10:57:40 Integrated Voltage Measurement... 10:57:49 Voltage Offset Correction: v_off_ini = 2475.169 uV, v_off_fin = 2473.016 uV 10:57:49 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -158.723 uVs 10:57:49 Integrated Voltage Measurement... 10:57:58 Voltage Offset Correction: v_off_ini = 2474.575 uV, v_off_fin = 2488.971 uV 10:57:58 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 175.346 uVs 10:58:02 I1Y Average and RMS of Measurements VT = -154.449 +- 11.296 uVs I1Y = -30.887 +- 2.260 uTm 10:58:07 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:58:10 Integrated Voltage Measurement... 10:58:19 Voltage Offset Correction: v_off_ini = 2472.993 uV, v_off_fin = 2457.719 uV 10:58:19 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -72.765 uVs 10:58:19 Integrated Voltage Measurement... 10:58:28 Voltage Offset Correction: v_off_ini = 2457.478 uV, v_off_fin = 2466.528 uV 10:58:28 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 104.534 uVs 10:58:28 Integrated Voltage Measurement... 10:58:37 Voltage Offset Correction: v_off_ini = 2458.829 uV, v_off_fin = 2460.146 uV 10:58:37 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -82.609 uVs 10:58:37 Integrated Voltage Measurement... 10:58:46 Voltage Offset Correction: v_off_ini = 2447.792 uV, v_off_fin = 2455.766 uV 10:58:46 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 86.256 uVs 10:58:46 Integrated Voltage Measurement... 10:58:55 Voltage Offset Correction: v_off_ini = 2454.940 uV, v_off_fin = 2466.527 uV 10:58:55 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -77.955 uVs 10:58:55 Integrated Voltage Measurement... 10:59:04 Voltage Offset Correction: v_off_ini = 2484.676 uV, v_off_fin = 2487.601 uV 10:59:04 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 115.589 uVs 10:59:06 I2X Average and RMS of Measurements VT = -89.951 +- 5.121 uVs I2X = 8.276 +- 0.471 uTm^2 10:59:12 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:59:14 Integrated Voltage Measurement... 10:59:23 Voltage Offset Correction: v_off_ini = 2495.925 uV, v_off_fin = 2476.875 uV 10:59:23 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -157.810 uVs 10:59:23 Integrated Voltage Measurement... 10:59:32 Voltage Offset Correction: v_off_ini = 2484.096 uV, v_off_fin = 2481.651 uV 10:59:32 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 177.292 uVs 10:59:32 Integrated Voltage Measurement... 10:59:41 Voltage Offset Correction: v_off_ini = 2459.349 uV, v_off_fin = 2459.218 uV 10:59:41 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -170.247 uVs 10:59:41 Integrated Voltage Measurement... 10:59:50 Voltage Offset Correction: v_off_ini = 2475.953 uV, v_off_fin = 2491.590 uV 10:59:50 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 148.555 uVs 10:59:50 Integrated Voltage Measurement... 10:59:59 Voltage Offset Correction: v_off_ini = 2465.392 uV, v_off_fin = 2470.619 uV 10:59:59 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -160.547 uVs 10:59:59 Integrated Voltage Measurement... 11:00:08 Voltage Offset Correction: v_off_ini = 2470.223 uV, v_off_fin = 2488.246 uV 11:00:08 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 162.369 uVs 11:00:11 I2Y Average and RMS of Measurements VT = -162.803 +- 3.461 uVs I2Y = -14.978 +- 0.318 uTm^2 11:00:21 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:00:26 Integrated Voltage Measurement... 11:00:35 Voltage Offset Correction: v_off_ini = 2513.022 uV, v_off_fin = 2501.893 uV 11:00:35 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -79.033 uVs 11:00:35 Integrated Voltage Measurement... 11:00:44 Voltage Offset Correction: v_off_ini = 2508.889 uV, v_off_fin = 2521.631 uV 11:00:44 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 100.446 uVs 11:00:44 Integrated Voltage Measurement... 11:00:53 Voltage Offset Correction: v_off_ini = 2527.126 uV, v_off_fin = 2518.410 uV 11:00:53 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -129.017 uVs 11:00:53 Integrated Voltage Measurement... 11:01:02 Voltage Offset Correction: v_off_ini = 2527.632 uV, v_off_fin = 2554.465 uV 11:01:02 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 90.768 uVs 11:01:02 Integrated Voltage Measurement... 11:01:11 Voltage Offset Correction: v_off_ini = 2551.407 uV, v_off_fin = 2559.374 uV 11:01:11 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -94.721 uVs 11:01:11 Integrated Voltage Measurement... 11:01:20 Voltage Offset Correction: v_off_ini = 2557.368 uV, v_off_fin = 2566.404 uV 11:01:20 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = 86.895 uVs 11:01:24 I1X Average and RMS of Measurements VT = -96.813 +- 9.259 uVs I1X = 19.363 +- 1.851 uTm 11:01:26 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:01:31 Integrated Voltage Measurement... 11:01:40 Voltage Offset Correction: v_off_ini = 2558.500 uV, v_off_fin = 2564.064 uV 11:01:40 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -153.302 uVs 11:01:40 Integrated Voltage Measurement... 11:01:49 Voltage Offset Correction: v_off_ini = 2560.533 uV, v_off_fin = 2565.790 uV 11:01:49 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 166.058 uVs 11:01:49 Integrated Voltage Measurement... 11:01:58 Voltage Offset Correction: v_off_ini = 2569.391 uV, v_off_fin = 2551.237 uV 11:01:58 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -173.520 uVs 11:01:58 Integrated Voltage Measurement... 11:02:07 Voltage Offset Correction: v_off_ini = 2537.125 uV, v_off_fin = 2559.630 uV 11:02:07 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5002 mm dXY = -0.5001 mm VT = 186.909 uVs 11:02:07 Integrated Voltage Measurement... 11:02:16 Voltage Offset Correction: v_off_ini = 2545.604 uV, v_off_fin = 2537.722 uV 11:02:16 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5002 mm dXY = 0.5001 mm VT = -161.553 uVs 11:02:16 Integrated Voltage Measurement... 11:02:25 Voltage Offset Correction: v_off_ini = 2562.278 uV, v_off_fin = 2589.971 uV 11:02:25 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 162.719 uVs 11:02:29 I1Y Average and RMS of Measurements VT = -167.344 +- 9.156 uVs I1Y = -33.465 +- 1.830 uTm 11:02:34 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:02:36 Integrated Voltage Measurement... 11:02:45 Voltage Offset Correction: v_off_ini = 2572.890 uV, v_off_fin = 2566.879 uV 11:02:45 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -78.492 uVs 11:02:46 Integrated Voltage Measurement... 11:02:54 Voltage Offset Correction: v_off_ini = 2552.523 uV, v_off_fin = 2551.126 uV 11:02:54 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 76.129 uVs 11:02:55 Integrated Voltage Measurement... 11:03:04 Voltage Offset Correction: v_off_ini = 2546.594 uV, v_off_fin = 2544.646 uV 11:03:04 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -82.177 uVs 11:03:04 Integrated Voltage Measurement... 11:03:13 Voltage Offset Correction: v_off_ini = 2548.573 uV, v_off_fin = 2564.410 uV 11:03:13 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 69.458 uVs 11:03:13 Integrated Voltage Measurement... 11:03:22 Voltage Offset Correction: v_off_ini = 2566.300 uV, v_off_fin = 2558.547 uV 11:03:22 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -93.294 uVs 11:03:22 Integrated Voltage Measurement... 11:03:31 Voltage Offset Correction: v_off_ini = 2558.542 uV, v_off_fin = 2564.250 uV 11:03:31 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 112.775 uVs 11:03:33 I2X Average and RMS of Measurements VT = -85.387 +- 12.493 uVs I2X = 7.856 +- 1.149 uTm^2 11:03:38 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:03:41 Integrated Voltage Measurement... 11:03:50 Voltage Offset Correction: v_off_ini = 2563.516 uV, v_off_fin = 2559.849 uV 11:03:50 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -187.138 uVs 11:03:50 Integrated Voltage Measurement... 11:03:59 Voltage Offset Correction: v_off_ini = 2566.579 uV, v_off_fin = 2549.165 uV 11:03:59 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 168.787 uVs 11:03:59 Integrated Voltage Measurement... 11:04:08 Voltage Offset Correction: v_off_ini = 2543.288 uV, v_off_fin = 2544.902 uV 11:04:08 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -135.595 uVs 11:04:08 Integrated Voltage Measurement... 11:04:17 Voltage Offset Correction: v_off_ini = 2551.154 uV, v_off_fin = 2564.927 uV 11:04:17 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 168.651 uVs 11:04:17 Integrated Voltage Measurement... 11:04:26 Voltage Offset Correction: v_off_ini = 2551.331 uV, v_off_fin = 2546.248 uV 11:04:26 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -180.127 uVs 11:04:26 Integrated Voltage Measurement... 11:04:35 Voltage Offset Correction: v_off_ini = 2537.289 uV, v_off_fin = 2556.113 uV 11:04:35 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 148.338 uVs 11:04:37 I2Y Average and RMS of Measurements VT = -164.773 +- 10.556 uVs I2Y = -15.159 +- 0.971 uTm^2 11:04:49 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:04:53 Integrated Voltage Measurement... 11:05:02 Voltage Offset Correction: v_off_ini = 2557.818 uV, v_off_fin = 2554.262 uV 11:05:02 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -101.574 uVs 11:05:02 Integrated Voltage Measurement... 11:05:11 Voltage Offset Correction: v_off_ini = 2595.485 uV, v_off_fin = 2586.018 uV 11:05:11 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 87.629 uVs 11:05:11 Integrated Voltage Measurement... 11:05:20 Voltage Offset Correction: v_off_ini = 2584.532 uV, v_off_fin = 2604.097 uV 11:05:20 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -79.041 uVs 11:05:20 Integrated Voltage Measurement... 11:05:29 Voltage Offset Correction: v_off_ini = 2614.516 uV, v_off_fin = 2606.231 uV 11:05:29 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 110.616 uVs 11:05:29 Integrated Voltage Measurement... 11:05:38 Voltage Offset Correction: v_off_ini = 2588.623 uV, v_off_fin = 2585.344 uV 11:05:38 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -105.034 uVs 11:05:38 Integrated Voltage Measurement... 11:05:47 Voltage Offset Correction: v_off_ini = 2594.882 uV, v_off_fin = 2608.427 uV 11:05:47 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 99.210 uVs 11:05:52 I1X Average and RMS of Measurements VT = -97.184 +- 3.493 uVs I1X = 19.437 +- 0.699 uTm 11:05:54 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:05:58 Integrated Voltage Measurement... 11:06:07 Voltage Offset Correction: v_off_ini = 2576.235 uV, v_off_fin = 2589.096 uV 11:06:07 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -152.036 uVs 11:06:07 Integrated Voltage Measurement... 11:06:16 Voltage Offset Correction: v_off_ini = 2593.541 uV, v_off_fin = 2614.398 uV 11:06:16 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 164.637 uVs 11:06:16 Integrated Voltage Measurement... 11:06:25 Voltage Offset Correction: v_off_ini = 2599.021 uV, v_off_fin = 2567.491 uV 11:06:25 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -152.726 uVs 11:06:25 Integrated Voltage Measurement... 11:06:34 Voltage Offset Correction: v_off_ini = 2581.478 uV, v_off_fin = 2612.733 uV 11:06:34 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 164.627 uVs 11:06:34 Integrated Voltage Measurement... 11:06:43 Voltage Offset Correction: v_off_ini = 2580.126 uV, v_off_fin = 2574.086 uV 11:06:43 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -153.127 uVs 11:06:43 Integrated Voltage Measurement... 11:06:52 Voltage Offset Correction: v_off_ini = 2608.616 uV, v_off_fin = 2597.892 uV 11:06:52 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 148.666 uVs 11:06:56 I1Y Average and RMS of Measurements VT = -155.970 +- 3.590 uVs I1Y = -31.192 +- 0.718 uTm 11:07:01 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:07:04 Integrated Voltage Measurement... 11:07:13 Voltage Offset Correction: v_off_ini = 2569.235 uV, v_off_fin = 2581.748 uV 11:07:13 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -72.071 uVs 11:07:13 Integrated Voltage Measurement... 11:07:22 Voltage Offset Correction: v_off_ini = 2595.811 uV, v_off_fin = 2584.276 uV 11:07:22 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 55.546 uVs 11:07:22 Integrated Voltage Measurement... 11:07:31 Voltage Offset Correction: v_off_ini = 2574.106 uV, v_off_fin = 2565.675 uV 11:07:31 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -83.105 uVs 11:07:31 Integrated Voltage Measurement... 11:07:40 Voltage Offset Correction: v_off_ini = 2585.688 uV, v_off_fin = 2590.791 uV 11:07:40 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 75.479 uVs 11:07:40 Integrated Voltage Measurement... 11:07:49 Voltage Offset Correction: v_off_ini = 2572.146 uV, v_off_fin = 2574.587 uV 11:07:49 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -83.396 uVs 11:07:49 Integrated Voltage Measurement... 11:07:58 Voltage Offset Correction: v_off_ini = 2585.794 uV, v_off_fin = 2615.295 uV 11:07:58 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 83.298 uVs 11:08:01 I2X Average and RMS of Measurements VT = -75.482 +- 8.419 uVs I2X = 6.944 +- 0.775 uTm^2 11:08:06 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:08:08 Integrated Voltage Measurement... 11:08:17 Voltage Offset Correction: v_off_ini = 2597.309 uV, v_off_fin = 2595.173 uV 11:08:17 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -173.483 uVs 11:08:17 Integrated Voltage Measurement... 11:08:26 Voltage Offset Correction: v_off_ini = 2611.974 uV, v_off_fin = 2604.497 uV 11:08:26 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 157.997 uVs 11:08:26 Integrated Voltage Measurement... 11:08:35 Voltage Offset Correction: v_off_ini = 2601.957 uV, v_off_fin = 2605.606 uV 11:08:35 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -152.172 uVs 11:08:35 Integrated Voltage Measurement... 11:08:44 Voltage Offset Correction: v_off_ini = 2595.336 uV, v_off_fin = 2590.626 uV 11:08:44 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 134.995 uVs 11:08:44 Integrated Voltage Measurement... 11:08:53 Voltage Offset Correction: v_off_ini = 2593.273 uV, v_off_fin = 2583.262 uV 11:08:53 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -171.660 uVs 11:08:53 Integrated Voltage Measurement... 11:09:02 Voltage Offset Correction: v_off_ini = 2586.242 uV, v_off_fin = 2589.032 uV 11:09:02 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 148.116 uVs 11:09:05 I2Y Average and RMS of Measurements VT = -156.404 +- 9.375 uVs I2Y = -14.389 +- 0.862 uTm^2 11:09:17 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:09:21 Integrated Voltage Measurement... 11:09:30 Voltage Offset Correction: v_off_ini = 2583.719 uV, v_off_fin = 2585.997 uV 11:09:30 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -88.015 uVs 11:09:30 Integrated Voltage Measurement... 11:09:39 Voltage Offset Correction: v_off_ini = 2567.817 uV, v_off_fin = 2576.181 uV 11:09:39 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 84.778 uVs 11:09:39 Integrated Voltage Measurement... 11:09:48 Voltage Offset Correction: v_off_ini = 2574.524 uV, v_off_fin = 2576.579 uV 11:09:48 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -62.658 uVs 11:09:48 Integrated Voltage Measurement... 11:09:57 Voltage Offset Correction: v_off_ini = 2559.224 uV, v_off_fin = 2553.747 uV 11:09:57 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 105.442 uVs 11:09:57 Integrated Voltage Measurement... 11:10:06 Voltage Offset Correction: v_off_ini = 2583.535 uV, v_off_fin = 2571.459 uV 11:10:06 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -92.258 uVs 11:10:06 Integrated Voltage Measurement... 11:10:15 Voltage Offset Correction: v_off_ini = 2582.713 uV, v_off_fin = 2574.202 uV 11:10:15 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 104.931 uVs 11:10:19 I1X Average and RMS of Measurements VT = -89.680 +- 6.376 uVs I1X = 17.937 +- 1.275 uTm 11:10:22 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:10:26 Integrated Voltage Measurement... 11:10:35 Voltage Offset Correction: v_off_ini = 2578.209 uV, v_off_fin = 2567.698 uV 11:10:35 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -149.388 uVs 11:10:35 Integrated Voltage Measurement... 11:10:44 Voltage Offset Correction: v_off_ini = 2591.590 uV, v_off_fin = 2610.882 uV 11:10:44 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 156.109 uVs 11:10:44 Integrated Voltage Measurement... 11:10:53 Voltage Offset Correction: v_off_ini = 2593.884 uV, v_off_fin = 2576.261 uV 11:10:53 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -150.730 uVs 11:10:53 Integrated Voltage Measurement... 11:11:02 Voltage Offset Correction: v_off_ini = 2581.762 uV, v_off_fin = 2609.883 uV 11:11:02 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 145.013 uVs 11:11:02 Integrated Voltage Measurement... 11:11:11 Voltage Offset Correction: v_off_ini = 2609.844 uV, v_off_fin = 2598.938 uV 11:11:11 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -127.939 uVs 11:11:11 Integrated Voltage Measurement... 11:11:20 Voltage Offset Correction: v_off_ini = 2576.910 uV, v_off_fin = 2585.670 uV 11:11:20 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 146.808 uVs 11:11:24 I1Y Average and RMS of Measurements VT = -145.998 +- 6.415 uVs I1Y = -29.198 +- 1.283 uTm 11:11:29 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:11:32 Integrated Voltage Measurement... 11:11:41 Voltage Offset Correction: v_off_ini = 2577.158 uV, v_off_fin = 2583.210 uV 11:11:41 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -84.523 uVs 11:11:41 Integrated Voltage Measurement... 11:11:50 Voltage Offset Correction: v_off_ini = 2567.870 uV, v_off_fin = 2559.871 uV 11:11:50 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 71.576 uVs 11:11:50 Integrated Voltage Measurement... 11:11:59 Voltage Offset Correction: v_off_ini = 2587.014 uV, v_off_fin = 2576.173 uV 11:11:59 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -82.576 uVs 11:11:59 Integrated Voltage Measurement... 11:12:08 Voltage Offset Correction: v_off_ini = 2589.829 uV, v_off_fin = 2595.352 uV 11:12:08 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 88.056 uVs 11:12:08 Integrated Voltage Measurement... 11:12:17 Voltage Offset Correction: v_off_ini = 2577.566 uV, v_off_fin = 2577.194 uV 11:12:17 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -77.244 uVs 11:12:17 Integrated Voltage Measurement... 11:12:26 Voltage Offset Correction: v_off_ini = 2581.512 uV, v_off_fin = 2576.162 uV 11:12:26 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 71.105 uVs 11:12:28 I2X Average and RMS of Measurements VT = -79.180 +- 4.618 uVs I2X = 7.285 +- 0.425 uTm^2 11:12:33 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:12:36 Integrated Voltage Measurement... 11:12:45 Voltage Offset Correction: v_off_ini = 2595.548 uV, v_off_fin = 2592.852 uV 11:12:45 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -170.446 uVs 11:12:45 Integrated Voltage Measurement... 11:12:54 Voltage Offset Correction: v_off_ini = 2601.958 uV, v_off_fin = 2618.706 uV 11:12:54 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 147.177 uVs 11:12:54 Integrated Voltage Measurement... 11:13:03 Voltage Offset Correction: v_off_ini = 2636.761 uV, v_off_fin = 2623.050 uV 11:13:03 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -134.624 uVs 11:13:03 Integrated Voltage Measurement... 11:13:12 Voltage Offset Correction: v_off_ini = 2597.647 uV, v_off_fin = 2611.181 uV 11:13:12 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 178.234 uVs 11:13:12 Integrated Voltage Measurement... 11:13:21 Voltage Offset Correction: v_off_ini = 2615.095 uV, v_off_fin = 2584.565 uV 11:13:21 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -194.856 uVs 11:13:21 Integrated Voltage Measurement... 11:13:30 Voltage Offset Correction: v_off_ini = 2598.534 uV, v_off_fin = 2605.169 uV 11:13:30 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 154.359 uVs 11:13:33 I2Y Average and RMS of Measurements VT = -163.283 +- 8.067 uVs I2Y = -15.022 +- 0.742 uTm^2 11:13:47 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:13:51 Integrated Voltage Measurement... 11:14:00 Voltage Offset Correction: v_off_ini = 2602.547 uV, v_off_fin = 2604.557 uV 11:14:00 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -75.443 uVs 11:14:00 Integrated Voltage Measurement... 11:14:09 Voltage Offset Correction: v_off_ini = 2618.790 uV, v_off_fin = 2613.817 uV 11:14:09 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 60.135 uVs 11:14:09 Integrated Voltage Measurement... 11:14:18 Voltage Offset Correction: v_off_ini = 2621.295 uV, v_off_fin = 2595.135 uV 11:14:18 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -56.818 uVs 11:14:18 Integrated Voltage Measurement... 11:14:27 Voltage Offset Correction: v_off_ini = 2615.811 uV, v_off_fin = 2645.853 uV 11:14:27 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 86.813 uVs 11:14:27 Integrated Voltage Measurement... 11:14:36 Voltage Offset Correction: v_off_ini = 2652.161 uV, v_off_fin = 2628.327 uV 11:14:36 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -83.189 uVs 11:14:36 Integrated Voltage Measurement... 11:14:45 Voltage Offset Correction: v_off_ini = 2644.993 uV, v_off_fin = 2655.518 uV 11:14:45 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 108.476 uVs 11:14:49 I1X Average and RMS of Measurements VT = -78.479 +- 12.380 uVs I1X = 15.696 +- 2.476 uTm 11:14:52 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:14:56 Integrated Voltage Measurement... 11:15:05 Voltage Offset Correction: v_off_ini = 2650.665 uV, v_off_fin = 2663.846 uV 11:15:05 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -142.449 uVs 11:15:05 Integrated Voltage Measurement... 11:15:14 Voltage Offset Correction: v_off_ini = 2669.850 uV, v_off_fin = 2682.885 uV 11:15:14 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 135.952 uVs 11:15:14 Integrated Voltage Measurement... 11:15:23 Voltage Offset Correction: v_off_ini = 2666.512 uV, v_off_fin = 2669.281 uV 11:15:23 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -140.046 uVs 11:15:23 Integrated Voltage Measurement... 11:15:32 Voltage Offset Correction: v_off_ini = 2646.761 uV, v_off_fin = 2680.524 uV 11:15:32 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 138.497 uVs 11:15:32 Integrated Voltage Measurement... 11:15:41 Voltage Offset Correction: v_off_ini = 2663.505 uV, v_off_fin = 2640.618 uV 11:15:41 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -147.307 uVs 11:15:41 Integrated Voltage Measurement... 11:15:50 Voltage Offset Correction: v_off_ini = 2661.490 uV, v_off_fin = 2662.262 uV 11:15:50 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 131.400 uVs 11:15:54 I1Y Average and RMS of Measurements VT = -139.275 +- 0.062 uVs I1Y = -27.853 +- 0.011 uTm 11:15:59 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:16:02 Integrated Voltage Measurement... 11:16:11 Voltage Offset Correction: v_off_ini = 2671.507 uV, v_off_fin = 2671.617 uV 11:16:11 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -58.941 uVs 11:16:11 Integrated Voltage Measurement... 11:16:20 Voltage Offset Correction: v_off_ini = 2687.921 uV, v_off_fin = 2682.657 uV 11:16:20 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 87.605 uVs 11:16:20 Integrated Voltage Measurement... 11:16:29 Voltage Offset Correction: v_off_ini = 2667.330 uV, v_off_fin = 2661.893 uV 11:16:29 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -76.197 uVs 11:16:29 Integrated Voltage Measurement... 11:16:38 Voltage Offset Correction: v_off_ini = 2684.474 uV, v_off_fin = 2695.787 uV 11:16:38 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 55.784 uVs 11:16:38 Integrated Voltage Measurement... 11:16:47 Voltage Offset Correction: v_off_ini = 2668.573 uV, v_off_fin = 2645.867 uV 11:16:47 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -71.110 uVs 11:16:47 Integrated Voltage Measurement... 11:16:56 Voltage Offset Correction: v_off_ini = 2665.512 uV, v_off_fin = 2694.459 uV 11:16:56 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 45.830 uVs 11:16:58 I2X Average and RMS of Measurements VT = -65.911 +- 6.044 uVs I2X = 6.064 +- 0.556 uTm^2 11:17:03 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:17:06 Integrated Voltage Measurement... 11:17:15 Voltage Offset Correction: v_off_ini = 2689.276 uV, v_off_fin = 2692.159 uV 11:17:15 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -161.049 uVs 11:17:15 Integrated Voltage Measurement... 11:17:24 Voltage Offset Correction: v_off_ini = 2712.993 uV, v_off_fin = 2694.964 uV 11:17:24 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 146.491 uVs 11:17:24 Integrated Voltage Measurement... 11:17:33 Voltage Offset Correction: v_off_ini = 2686.937 uV, v_off_fin = 2666.556 uV 11:17:33 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -128.029 uVs 11:17:33 Integrated Voltage Measurement... 11:17:42 Voltage Offset Correction: v_off_ini = 2663.268 uV, v_off_fin = 2653.051 uV 11:17:42 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 157.816 uVs 11:17:42 Integrated Voltage Measurement... 11:17:51 Voltage Offset Correction: v_off_ini = 2678.000 uV, v_off_fin = 2680.718 uV 11:17:51 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -134.007 uVs 11:17:51 Integrated Voltage Measurement... 11:18:00 Voltage Offset Correction: v_off_ini = 2695.389 uV, v_off_fin = 2705.589 uV 11:18:00 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 142.169 uVs 11:18:02 I2Y Average and RMS of Measurements VT = -144.927 +- 6.557 uVs I2Y = -13.333 +- 0.603 uTm^2 11:18:16 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:18:21 Integrated Voltage Measurement... 11:18:30 Voltage Offset Correction: v_off_ini = 2736.846 uV, v_off_fin = 2721.618 uV 11:18:30 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -61.266 uVs 11:18:30 Integrated Voltage Measurement... 11:18:39 Voltage Offset Correction: v_off_ini = 2709.160 uV, v_off_fin = 2704.633 uV 11:18:39 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 60.603 uVs 11:18:39 Integrated Voltage Measurement... 11:18:48 Voltage Offset Correction: v_off_ini = 2726.052 uV, v_off_fin = 2723.000 uV 11:18:48 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -71.087 uVs 11:18:48 Integrated Voltage Measurement... 11:18:56 Voltage Offset Correction: v_off_ini = 2728.454 uV, v_off_fin = 2728.169 uV 11:18:57 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 70.112 uVs 11:18:57 Integrated Voltage Measurement... 11:19:05 Voltage Offset Correction: v_off_ini = 2735.134 uV, v_off_fin = 2720.847 uV 11:19:06 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -97.035 uVs 11:19:06 Integrated Voltage Measurement... 11:19:14 Voltage Offset Correction: v_off_ini = 2729.823 uV, v_off_fin = 2745.230 uV 11:19:15 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 100.102 uVs 11:19:19 I1X Average and RMS of Measurements VT = -76.701 +- 15.958 uVs I1X = 15.340 +- 3.192 uTm 11:19:21 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:19:25 Integrated Voltage Measurement... 11:19:34 Voltage Offset Correction: v_off_ini = 2728.860 uV, v_off_fin = 2711.182 uV 11:19:34 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -129.834 uVs 11:19:34 Integrated Voltage Measurement... 11:19:43 Voltage Offset Correction: v_off_ini = 2735.268 uV, v_off_fin = 2743.962 uV 11:19:43 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 168.633 uVs 11:19:43 Integrated Voltage Measurement... 11:19:52 Voltage Offset Correction: v_off_ini = 2737.577 uV, v_off_fin = 2728.582 uV 11:19:52 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5002 mm dXY = 0.5001 mm VT = -133.484 uVs 11:19:52 Integrated Voltage Measurement... 11:20:01 Voltage Offset Correction: v_off_ini = 2725.430 uV, v_off_fin = 2724.642 uV 11:20:01 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5002 mm dXY = -0.5001 mm VT = 135.899 uVs 11:20:01 Integrated Voltage Measurement... 11:20:10 Voltage Offset Correction: v_off_ini = 2726.425 uV, v_off_fin = 2697.915 uV 11:20:10 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -145.909 uVs 11:20:10 Integrated Voltage Measurement... 11:20:19 Voltage Offset Correction: v_off_ini = 2712.155 uV, v_off_fin = 2757.153 uV 11:20:19 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 185.759 uVs 11:20:24 I1Y Average and RMS of Measurements VT = -149.920 +- 12.723 uVs I1Y = -29.980 +- 2.545 uTm 11:20:29 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:20:31 Integrated Voltage Measurement... 11:20:40 Voltage Offset Correction: v_off_ini = 2735.086 uV, v_off_fin = 2727.763 uV 11:20:40 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -43.681 uVs 11:20:40 Integrated Voltage Measurement... 11:20:49 Voltage Offset Correction: v_off_ini = 2733.122 uV, v_off_fin = 2737.090 uV 11:20:49 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 67.017 uVs 11:20:49 Integrated Voltage Measurement... 11:20:58 Voltage Offset Correction: v_off_ini = 2749.515 uV, v_off_fin = 2737.776 uV 11:20:58 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -70.775 uVs 11:20:58 Integrated Voltage Measurement... 11:21:07 Voltage Offset Correction: v_off_ini = 2743.438 uV, v_off_fin = 2761.222 uV 11:21:07 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 79.977 uVs 11:21:07 Integrated Voltage Measurement... 11:21:16 Voltage Offset Correction: v_off_ini = 2730.263 uV, v_off_fin = 2739.382 uV 11:21:16 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -36.440 uVs 11:21:16 Integrated Voltage Measurement... 11:21:25 Voltage Offset Correction: v_off_ini = 2748.333 uV, v_off_fin = 2761.477 uV 11:21:25 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 68.198 uVs 11:21:28 I2X Average and RMS of Measurements VT = -61.015 +- 10.230 uVs I2X = 5.613 +- 0.941 uTm^2 11:21:33 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:21:36 Integrated Voltage Measurement... 11:21:45 Voltage Offset Correction: v_off_ini = 2769.047 uV, v_off_fin = 2749.414 uV 11:21:45 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -170.257 uVs 11:21:45 Integrated Voltage Measurement... 11:21:54 Voltage Offset Correction: v_off_ini = 2777.668 uV, v_off_fin = 2774.117 uV 11:21:54 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 144.464 uVs 11:21:54 Integrated Voltage Measurement... 11:22:03 Voltage Offset Correction: v_off_ini = 2771.230 uV, v_off_fin = 2784.701 uV 11:22:03 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -131.815 uVs 11:22:03 Integrated Voltage Measurement... 11:22:12 Voltage Offset Correction: v_off_ini = 2806.836 uV, v_off_fin = 2794.084 uV 11:22:12 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 155.585 uVs 11:22:12 Integrated Voltage Measurement... 11:22:21 Voltage Offset Correction: v_off_ini = 2768.347 uV, v_off_fin = 2790.161 uV 11:22:21 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -133.999 uVs 11:22:21 Integrated Voltage Measurement... 11:22:30 Voltage Offset Correction: v_off_ini = 2783.203 uV, v_off_fin = 2788.552 uV 11:22:30 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 144.704 uVs 11:22:32 I2Y Average and RMS of Measurements VT = -146.804 +- 7.673 uVs I2Y = -13.506 +- 0.706 uTm^2 11:22:46 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:22:50 Integrated Voltage Measurement... 11:22:59 Voltage Offset Correction: v_off_ini = 2794.527 uV, v_off_fin = 2782.419 uV 11:22:59 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -97.505 uVs 11:22:59 Integrated Voltage Measurement... 11:23:08 Voltage Offset Correction: v_off_ini = 2802.302 uV, v_off_fin = 2793.323 uV 11:23:08 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 66.784 uVs 11:23:08 Integrated Voltage Measurement... 11:23:17 Voltage Offset Correction: v_off_ini = 2812.071 uV, v_off_fin = 2782.335 uV 11:23:17 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -88.155 uVs 11:23:18 Integrated Voltage Measurement... 11:23:26 Voltage Offset Correction: v_off_ini = 2816.051 uV, v_off_fin = 2822.165 uV 11:23:26 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 62.279 uVs 11:23:27 Integrated Voltage Measurement... 11:23:35 Voltage Offset Correction: v_off_ini = 2798.395 uV, v_off_fin = 2801.786 uV 11:23:36 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -80.455 uVs 11:23:36 Integrated Voltage Measurement... 11:23:44 Voltage Offset Correction: v_off_ini = 2782.551 uV, v_off_fin = 2795.667 uV 11:23:45 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 70.968 uVs 11:23:49 I1X Average and RMS of Measurements VT = -77.691 +- 3.155 uVs I1X = 15.539 +- 0.631 uTm 11:23:51 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:23:55 Integrated Voltage Measurement... 11:24:04 Voltage Offset Correction: v_off_ini = 2797.369 uV, v_off_fin = 2794.103 uV 11:24:04 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -134.314 uVs 11:24:04 Integrated Voltage Measurement... 11:24:13 Voltage Offset Correction: v_off_ini = 2814.165 uV, v_off_fin = 2815.321 uV 11:24:13 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 156.813 uVs 11:24:13 Integrated Voltage Measurement... 11:24:22 Voltage Offset Correction: v_off_ini = 2816.185 uV, v_off_fin = 2801.458 uV 11:24:22 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -169.754 uVs 11:24:22 Integrated Voltage Measurement... 11:24:31 Voltage Offset Correction: v_off_ini = 2795.327 uV, v_off_fin = 2791.411 uV 11:24:31 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5002 mm dXY = -0.5001 mm VT = 169.916 uVs 11:24:31 Integrated Voltage Measurement... 11:24:40 Voltage Offset Correction: v_off_ini = 2795.082 uV, v_off_fin = 2774.780 uV 11:24:40 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -141.651 uVs 11:24:40 Integrated Voltage Measurement... 11:24:49 Voltage Offset Correction: v_off_ini = 2777.880 uV, v_off_fin = 2788.833 uV 11:24:49 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 154.977 uVs 11:24:54 I1Y Average and RMS of Measurements VT = -154.571 +- 10.852 uVs I1Y = -30.911 +- 2.169 uTm 11:24:59 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:25:01 Integrated Voltage Measurement... 11:25:10 Voltage Offset Correction: v_off_ini = 2759.902 uV, v_off_fin = 2765.545 uV 11:25:10 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -38.108 uVs 11:25:10 Integrated Voltage Measurement... 11:25:19 Voltage Offset Correction: v_off_ini = 2781.455 uV, v_off_fin = 2785.847 uV 11:25:19 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 46.386 uVs 11:25:19 Integrated Voltage Measurement... 11:25:28 Voltage Offset Correction: v_off_ini = 2790.599 uV, v_off_fin = 2765.772 uV 11:25:28 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -75.164 uVs 11:25:28 Integrated Voltage Measurement... 11:25:37 Voltage Offset Correction: v_off_ini = 2808.144 uV, v_off_fin = 2803.680 uV 11:25:37 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 71.979 uVs 11:25:37 Integrated Voltage Measurement... 11:25:46 Voltage Offset Correction: v_off_ini = 2803.800 uV, v_off_fin = 2788.937 uV 11:25:46 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -74.244 uVs 11:25:46 Integrated Voltage Measurement... 11:25:55 Voltage Offset Correction: v_off_ini = 2786.762 uV, v_off_fin = 2804.506 uV 11:25:55 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 77.527 uVs 11:25:58 I2X Average and RMS of Measurements VT = -63.901 +- 15.341 uVs I2X = 5.879 +- 1.411 uTm^2 11:26:03 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:26:05 Integrated Voltage Measurement... 11:26:14 Voltage Offset Correction: v_off_ini = 2797.974 uV, v_off_fin = 2808.800 uV 11:26:15 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -163.788 uVs 11:26:15 Integrated Voltage Measurement... 11:26:23 Voltage Offset Correction: v_off_ini = 2808.244 uV, v_off_fin = 2828.784 uV 11:26:24 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 177.202 uVs 11:26:24 Integrated Voltage Measurement... 11:26:33 Voltage Offset Correction: v_off_ini = 2839.364 uV, v_off_fin = 2812.699 uV 11:26:33 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -146.778 uVs 11:26:33 Integrated Voltage Measurement... 11:26:42 Voltage Offset Correction: v_off_ini = 2822.983 uV, v_off_fin = 2819.469 uV 11:26:42 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 167.767 uVs 11:26:42 Integrated Voltage Measurement... 11:26:51 Voltage Offset Correction: v_off_ini = 2833.574 uV, v_off_fin = 2829.644 uV 11:26:51 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -116.812 uVs 11:26:51 Integrated Voltage Measurement... 11:27:00 Voltage Offset Correction: v_off_ini = 2847.153 uV, v_off_fin = 2835.571 uV 11:27:00 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 128.395 uVs 11:27:02 I2Y Average and RMS of Measurements VT = -150.124 +- 20.195 uVs I2Y = -13.811 +- 1.858 uTm^2 11:27:16 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:27:20 Integrated Voltage Measurement... 11:27:29 Voltage Offset Correction: v_off_ini = 2853.106 uV, v_off_fin = 2847.709 uV 11:27:29 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -73.614 uVs 11:27:29 Integrated Voltage Measurement... 11:27:38 Voltage Offset Correction: v_off_ini = 2855.366 uV, v_off_fin = 2844.011 uV 11:27:38 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 81.773 uVs 11:27:38 Integrated Voltage Measurement... 11:27:47 Voltage Offset Correction: v_off_ini = 2837.355 uV, v_off_fin = 2834.246 uV 11:27:47 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -92.826 uVs 11:27:47 Integrated Voltage Measurement... 11:27:56 Voltage Offset Correction: v_off_ini = 2833.681 uV, v_off_fin = 2872.963 uV 11:27:56 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 77.898 uVs 11:27:56 Integrated Voltage Measurement... 11:28:05 Voltage Offset Correction: v_off_ini = 2849.893 uV, v_off_fin = 2859.107 uV 11:28:05 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -69.896 uVs 11:28:05 Integrated Voltage Measurement... 11:28:14 Voltage Offset Correction: v_off_ini = 2862.368 uV, v_off_fin = 2852.895 uV 11:28:14 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 80.009 uVs 11:28:19 I1X Average and RMS of Measurements VT = -79.336 +- 4.406 uVs I1X = 15.867 +- 0.881 uTm 11:28:21 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:28:25 Integrated Voltage Measurement... 11:28:34 Voltage Offset Correction: v_off_ini = 2843.143 uV, v_off_fin = 2843.296 uV 11:28:34 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = -122.308 uVs 11:28:34 Integrated Voltage Measurement... 11:28:43 Voltage Offset Correction: v_off_ini = 2855.271 uV, v_off_fin = 2871.860 uV 11:28:43 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 150.117 uVs 11:28:43 Integrated Voltage Measurement... 11:28:52 Voltage Offset Correction: v_off_ini = 2849.271 uV, v_off_fin = 2861.429 uV 11:28:52 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -142.860 uVs 11:28:52 Integrated Voltage Measurement... 11:29:01 Voltage Offset Correction: v_off_ini = 2847.867 uV, v_off_fin = 2858.959 uV 11:29:01 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 171.040 uVs 11:29:01 Integrated Voltage Measurement... 11:29:10 Voltage Offset Correction: v_off_ini = 2853.801 uV, v_off_fin = 2832.534 uV 11:29:10 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -168.956 uVs 11:29:10 Integrated Voltage Measurement... 11:29:19 Voltage Offset Correction: v_off_ini = 2843.905 uV, v_off_fin = 2844.600 uV 11:29:19 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 141.956 uVs 11:29:23 I1Y Average and RMS of Measurements VT = -149.540 +- 9.443 uVs I1Y = -29.907 +- 1.887 uTm 11:29:28 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:29:31 Integrated Voltage Measurement... 11:29:40 Voltage Offset Correction: v_off_ini = 2869.616 uV, v_off_fin = 2858.688 uV 11:29:40 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -69.259 uVs 11:29:40 Integrated Voltage Measurement... 11:29:49 Voltage Offset Correction: v_off_ini = 2847.329 uV, v_off_fin = 2865.218 uV 11:29:49 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 69.689 uVs 11:29:49 Integrated Voltage Measurement... 11:29:58 Voltage Offset Correction: v_off_ini = 2853.496 uV, v_off_fin = 2851.533 uV 11:29:58 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -79.590 uVs 11:29:58 Integrated Voltage Measurement... 11:30:07 Voltage Offset Correction: v_off_ini = 2849.813 uV, v_off_fin = 2860.705 uV 11:30:07 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 73.746 uVs 11:30:07 Integrated Voltage Measurement... 11:30:16 Voltage Offset Correction: v_off_ini = 2839.229 uV, v_off_fin = 2836.565 uV 11:30:16 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -60.431 uVs 11:30:16 Integrated Voltage Measurement... 11:30:25 Voltage Offset Correction: v_off_ini = 2849.537 uV, v_off_fin = 2869.342 uV 11:30:25 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 78.107 uVs 11:30:27 I2X Average and RMS of Measurements VT = -71.804 +- 3.440 uVs I2X = 6.606 +- 0.317 uTm^2 11:30:32 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:30:35 Integrated Voltage Measurement... 11:30:44 Voltage Offset Correction: v_off_ini = 2880.054 uV, v_off_fin = 2878.342 uV 11:30:44 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -155.225 uVs 11:30:44 Integrated Voltage Measurement... 11:30:53 Voltage Offset Correction: v_off_ini = 2869.113 uV, v_off_fin = 2881.517 uV 11:30:53 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 140.807 uVs 11:30:53 Integrated Voltage Measurement... 11:31:02 Voltage Offset Correction: v_off_ini = 2877.209 uV, v_off_fin = 2865.351 uV 11:31:02 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -146.729 uVs 11:31:02 Integrated Voltage Measurement... 11:31:11 Voltage Offset Correction: v_off_ini = 2868.733 uV, v_off_fin = 2871.344 uV 11:31:11 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 193.684 uVs 11:31:11 Integrated Voltage Measurement... 11:31:20 Voltage Offset Correction: v_off_ini = 2858.770 uV, v_off_fin = 2877.040 uV 11:31:20 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -144.415 uVs 11:31:20 Integrated Voltage Measurement... 11:31:29 Voltage Offset Correction: v_off_ini = 2849.818 uV, v_off_fin = 2864.244 uV 11:31:29 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 168.359 uVs 11:31:32 I2Y Average and RMS of Measurements VT = -158.203 +- 9.150 uVs I2Y = -14.555 +- 0.842 uTm^2 11:31:46 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:31:50 Integrated Voltage Measurement... 11:31:59 Voltage Offset Correction: v_off_ini = 2892.706 uV, v_off_fin = 2863.719 uV 11:31:59 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -94.603 uVs 11:31:59 Integrated Voltage Measurement... 11:32:08 Voltage Offset Correction: v_off_ini = 2871.973 uV, v_off_fin = 2865.865 uV 11:32:08 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 70.805 uVs 11:32:08 Integrated Voltage Measurement... 11:32:17 Voltage Offset Correction: v_off_ini = 2880.352 uV, v_off_fin = 2871.143 uV 11:32:17 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -91.137 uVs 11:32:17 Integrated Voltage Measurement... 11:32:26 Voltage Offset Correction: v_off_ini = 2869.372 uV, v_off_fin = 2868.900 uV 11:32:26 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 54.786 uVs 11:32:26 Integrated Voltage Measurement... 11:32:35 Voltage Offset Correction: v_off_ini = 2876.649 uV, v_off_fin = 2881.397 uV 11:32:35 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -80.125 uVs 11:32:35 Integrated Voltage Measurement... 11:32:44 Voltage Offset Correction: v_off_ini = 2892.024 uV, v_off_fin = 2887.083 uV 11:32:44 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 97.373 uVs 11:32:48 I1X Average and RMS of Measurements VT = -81.472 +- 6.504 uVs I1X = 16.294 +- 1.301 uTm 11:32:51 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:32:55 Integrated Voltage Measurement... 11:33:04 Voltage Offset Correction: v_off_ini = 2844.924 uV, v_off_fin = 2845.472 uV 11:33:04 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -122.506 uVs 11:33:04 Integrated Voltage Measurement... 11:33:13 Voltage Offset Correction: v_off_ini = 2852.010 uV, v_off_fin = 2846.878 uV 11:33:13 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 167.327 uVs 11:33:13 Integrated Voltage Measurement... 11:33:22 Voltage Offset Correction: v_off_ini = 2869.516 uV, v_off_fin = 2860.253 uV 11:33:22 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -156.755 uVs 11:33:22 Integrated Voltage Measurement... 11:33:31 Voltage Offset Correction: v_off_ini = 2883.372 uV, v_off_fin = 2908.379 uV 11:33:31 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 154.855 uVs 11:33:31 Integrated Voltage Measurement... 11:33:40 Voltage Offset Correction: v_off_ini = 2882.881 uV, v_off_fin = 2858.707 uV 11:33:40 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -141.836 uVs 11:33:40 Integrated Voltage Measurement... 11:33:49 Voltage Offset Correction: v_off_ini = 2848.311 uV, v_off_fin = 2855.283 uV 11:33:49 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 164.996 uVs 11:33:53 I1Y Average and RMS of Measurements VT = -151.379 +- 4.673 uVs I1Y = -30.273 +- 0.934 uTm 11:33:58 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:34:01 Integrated Voltage Measurement... 11:34:09 Voltage Offset Correction: v_off_ini = 2839.666 uV, v_off_fin = 2876.983 uV 11:34:09 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -63.062 uVs 11:34:10 Integrated Voltage Measurement... 11:34:18 Voltage Offset Correction: v_off_ini = 2878.360 uV, v_off_fin = 2864.866 uV 11:34:18 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 72.955 uVs 11:34:19 Integrated Voltage Measurement... 11:34:27 Voltage Offset Correction: v_off_ini = 2836.579 uV, v_off_fin = 2824.606 uV 11:34:27 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -61.384 uVs 11:34:28 Integrated Voltage Measurement... 11:34:36 Voltage Offset Correction: v_off_ini = 2834.817 uV, v_off_fin = 2846.639 uV 11:34:37 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 76.984 uVs 11:34:37 Integrated Voltage Measurement... 11:34:45 Voltage Offset Correction: v_off_ini = 2839.652 uV, v_off_fin = 2838.113 uV 11:34:46 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -70.734 uVs 11:34:46 Integrated Voltage Measurement... 11:34:55 Voltage Offset Correction: v_off_ini = 2838.296 uV, v_off_fin = 2862.523 uV 11:34:55 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 70.987 uVs 11:34:57 I2X Average and RMS of Measurements VT = -69.351 +- 1.170 uVs I2X = 6.380 +- 0.108 uTm^2 11:35:02 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:35:05 Integrated Voltage Measurement... 11:35:14 Voltage Offset Correction: v_off_ini = 2849.734 uV, v_off_fin = 2828.979 uV 11:35:14 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -170.108 uVs 11:35:14 Integrated Voltage Measurement... 11:35:23 Voltage Offset Correction: v_off_ini = 2823.247 uV, v_off_fin = 2832.084 uV 11:35:23 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 167.122 uVs 11:35:23 Integrated Voltage Measurement... 11:35:32 Voltage Offset Correction: v_off_ini = 2818.812 uV, v_off_fin = 2828.944 uV 11:35:32 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -160.128 uVs 11:35:32 Integrated Voltage Measurement... 11:35:41 Voltage Offset Correction: v_off_ini = 2796.081 uV, v_off_fin = 2815.011 uV 11:35:41 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 170.282 uVs 11:35:41 Integrated Voltage Measurement... 11:35:50 Voltage Offset Correction: v_off_ini = 2800.650 uV, v_off_fin = 2794.302 uV 11:35:50 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -163.377 uVs 11:35:50 Integrated Voltage Measurement... 11:35:59 Voltage Offset Correction: v_off_ini = 2779.466 uV, v_off_fin = 2789.270 uV 11:35:59 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 167.818 uVs 11:36:01 I2Y Average and RMS of Measurements VT = -166.473 +- 1.523 uVs I2Y = -15.315 +- 0.140 uTm^2 11:36:15 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:36:19 Integrated Voltage Measurement... 11:36:28 Voltage Offset Correction: v_off_ini = 2779.117 uV, v_off_fin = 2787.216 uV 11:36:28 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -89.332 uVs 11:36:28 Integrated Voltage Measurement... 11:36:37 Voltage Offset Correction: v_off_ini = 2781.594 uV, v_off_fin = 2782.910 uV 11:36:37 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 77.931 uVs 11:36:37 Integrated Voltage Measurement... 11:36:46 Voltage Offset Correction: v_off_ini = 2768.227 uV, v_off_fin = 2765.529 uV 11:36:46 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -106.903 uVs 11:36:46 Integrated Voltage Measurement... 11:36:55 Voltage Offset Correction: v_off_ini = 2783.424 uV, v_off_fin = 2773.315 uV 11:36:55 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 101.493 uVs 11:36:55 Integrated Voltage Measurement... 11:37:04 Voltage Offset Correction: v_off_ini = 2795.462 uV, v_off_fin = 2790.866 uV 11:37:04 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -85.075 uVs 11:37:04 Integrated Voltage Measurement... 11:37:13 Voltage Offset Correction: v_off_ini = 2789.083 uV, v_off_fin = 2774.833 uV 11:37:13 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 86.233 uVs 11:37:17 I1X Average and RMS of Measurements VT = -91.161 +- 9.255 uVs I1X = 18.233 +- 1.852 uTm 11:37:20 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:37:24 Integrated Voltage Measurement... 11:37:33 Voltage Offset Correction: v_off_ini = 2787.217 uV, v_off_fin = 2769.848 uV 11:37:33 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = -160.144 uVs 11:37:33 Integrated Voltage Measurement... 11:37:42 Voltage Offset Correction: v_off_ini = 2778.318 uV, v_off_fin = 2793.499 uV 11:37:42 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = 172.824 uVs 11:37:42 Integrated Voltage Measurement... 11:37:51 Voltage Offset Correction: v_off_ini = 2782.397 uV, v_off_fin = 2779.701 uV 11:37:51 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -150.912 uVs 11:37:51 Integrated Voltage Measurement... 11:38:00 Voltage Offset Correction: v_off_ini = 2795.165 uV, v_off_fin = 2801.580 uV 11:38:00 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 191.406 uVs 11:38:00 Integrated Voltage Measurement... 11:38:09 Voltage Offset Correction: v_off_ini = 2804.598 uV, v_off_fin = 2787.458 uV 11:38:09 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -155.605 uVs 11:38:09 Integrated Voltage Measurement... 11:38:18 Voltage Offset Correction: v_off_ini = 2810.230 uV, v_off_fin = 2826.597 uV 11:38:18 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = 158.811 uVs 11:38:22 I1Y Average and RMS of Measurements VT = -164.950 +- 5.798 uVs I1Y = -32.991 +- 1.158 uTm 11:38:27 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:38:30 Integrated Voltage Measurement... 11:38:39 Voltage Offset Correction: v_off_ini = 2786.689 uV, v_off_fin = 2797.052 uV 11:38:39 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -80.739 uVs 11:38:39 Integrated Voltage Measurement... 11:38:48 Voltage Offset Correction: v_off_ini = 2804.032 uV, v_off_fin = 2836.837 uV 11:38:48 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 85.483 uVs 11:38:48 Integrated Voltage Measurement... 11:38:57 Voltage Offset Correction: v_off_ini = 2818.459 uV, v_off_fin = 2823.391 uV 11:38:57 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -73.230 uVs 11:38:57 Integrated Voltage Measurement... 11:39:06 Voltage Offset Correction: v_off_ini = 2825.805 uV, v_off_fin = 2842.579 uV 11:39:06 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 67.499 uVs 11:39:06 Integrated Voltage Measurement... 11:39:15 Voltage Offset Correction: v_off_ini = 2853.473 uV, v_off_fin = 2846.775 uV 11:39:15 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -71.292 uVs 11:39:15 Integrated Voltage Measurement... 11:39:24 Voltage Offset Correction: v_off_ini = 2852.795 uV, v_off_fin = 2869.927 uV 11:39:24 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 98.079 uVs 11:39:26 I2X Average and RMS of Measurements VT = -79.387 +- 6.412 uVs I2X = 7.304 +- 0.590 uTm^2 11:39:31 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:39:34 Integrated Voltage Measurement... 11:39:43 Voltage Offset Correction: v_off_ini = 2819.074 uV, v_off_fin = 2810.064 uV 11:39:43 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = -159.830 uVs 11:39:43 Integrated Voltage Measurement... 11:39:52 Voltage Offset Correction: v_off_ini = 2813.967 uV, v_off_fin = 2806.574 uV 11:39:52 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 155.380 uVs 11:39:52 Integrated Voltage Measurement... 11:40:01 Voltage Offset Correction: v_off_ini = 2842.020 uV, v_off_fin = 2818.928 uV 11:40:01 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -183.958 uVs 11:40:01 Integrated Voltage Measurement... 11:40:10 Voltage Offset Correction: v_off_ini = 2821.436 uV, v_off_fin = 2809.089 uV 11:40:10 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 182.492 uVs 11:40:10 Integrated Voltage Measurement... 11:40:19 Voltage Offset Correction: v_off_ini = 2828.094 uV, v_off_fin = 2834.520 uV 11:40:19 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -167.119 uVs 11:40:19 Integrated Voltage Measurement... 11:40:28 Voltage Offset Correction: v_off_ini = 2842.628 uV, v_off_fin = 2839.564 uV 11:40:28 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 155.940 uVs 11:40:30 I2Y Average and RMS of Measurements VT = -167.453 +- 11.267 uVs I2Y = -15.406 +- 1.036 uTm^2 11:40:33 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm