SLAC Magnetic Measurements Date: 08-24-2026 Time: 16:27:25 Project: LCLS-II-HE Magnet Type: Phase Shifter Magnet Name: HE_SXPS_15081 Dataset Number: 6 Run Purpose: Final Results Run Type: Run Number: 7 Measurement Device: Stretched Wire Operator: YL Comment: Off-Axis 16:27:34 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:27:36 Motion Parameters, X Controller, Axis 1 acc = 0.001000 m/s/s dcl = 0.001000 m/s/s vel = 0.001000 m/s pos = 0.000000 m kp = 188.600000 ki = 820.000000 kd = 491.400000 db = 0.000000 16:27:38 Motion Parameters, X Controller, Axis 2 acc = 0.001000 m/s/s dcl = 0.001000 m/s/s vel = 0.001000 m/s pos = 0.000000 m kp = 188.600000 ki = 820.000000 kd = 491.400000 db = 0.000000 16:27:40 Motion Parameters, Y Controller, Axis 1 acc = 0.001000 m/s/s dcl = 0.001000 m/s/s vel = 0.001000 m/s pos = 0.000000 m kp = 188.600000 ki = 820.000000 kd = 491.400000 db = 0.000000 16:27:41 Motion Parameters, Y Controller, Axis 2 acc = 0.001000 m/s/s dcl = 0.001000 m/s/s vel = 0.001000 m/s pos = 0.000000 m kp = 188.600000 ki = 820.000000 kd = 491.400000 db = 0.000000 16:27:59 Wire Position: X0 = -2.0000 mm, Y0 = 0.0000 mm 16:28:03 Integrated Voltage Measurement... 16:28:12 Voltage Offset Correction: v_off_ini = 1647.379 uV, v_off_fin = 1636.292 uV 16:28:12 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -196.585 uVs 16:28:12 Integrated Voltage Measurement... 16:28:21 Voltage Offset Correction: v_off_ini = 1619.562 uV, v_off_fin = 1614.254 uV 16:28:21 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = 167.375 uVs 16:28:21 Integrated Voltage Measurement... 16:28:30 Voltage Offset Correction: v_off_ini = 1617.488 uV, v_off_fin = 1613.337 uV 16:28:30 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -191.792 uVs 16:28:30 Integrated Voltage Measurement... 16:28:39 Voltage Offset Correction: v_off_ini = 1640.599 uV, v_off_fin = 1633.448 uV 16:28:39 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 186.640 uVs 16:28:39 Integrated Voltage Measurement... 16:28:48 Voltage Offset Correction: v_off_ini = 1614.242 uV, v_off_fin = 1595.955 uV 16:28:48 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -197.520 uVs 16:28:48 Integrated Voltage Measurement... 16:28:57 Voltage Offset Correction: v_off_ini = 1606.061 uV, v_off_fin = 1610.518 uV 16:28:57 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 179.254 uVs 16:29:01 I1X Average and RMS of Measurements VT = -186.528 +- 3.233 uVs I1X = 37.307 +- 0.646 uTm 16:29:04 Wire Position: X0 = -2.0000 mm, Y0 = 0.0000 mm 16:29:08 Integrated Voltage Measurement... 16:29:17 Voltage Offset Correction: v_off_ini = 1612.074 uV, v_off_fin = 1610.238 uV 16:29:17 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.4999 mm VT = -191.229 uVs 16:29:17 Integrated Voltage Measurement... 16:29:26 Voltage Offset Correction: v_off_ini = 1620.654 uV, v_off_fin = 1637.649 uV 16:29:26 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 219.290 uVs 16:29:26 Integrated Voltage Measurement... 16:29:35 Voltage Offset Correction: v_off_ini = 1606.822 uV, v_off_fin = 1581.979 uV 16:29:35 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -194.539 uVs 16:29:35 Integrated Voltage Measurement... 16:29:44 Voltage Offset Correction: v_off_ini = 1610.229 uV, v_off_fin = 1612.809 uV 16:29:44 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 203.277 uVs 16:29:44 Integrated Voltage Measurement... 16:29:53 Voltage Offset Correction: v_off_ini = 1624.902 uV, v_off_fin = 1619.550 uV 16:29:53 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -201.166 uVs 16:29:53 Integrated Voltage Measurement... 16:30:02 Voltage Offset Correction: v_off_ini = 1611.754 uV, v_off_fin = 1626.682 uV 16:30:02 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 204.554 uVs 16:30:06 I1Y Average and RMS of Measurements VT = -202.342 +- 2.619 uVs I1Y = -40.469 +- 0.524 uTm 16:30:11 Wire Position: X0 = -2.0000 mm, Y0 = 0.0000 mm 16:30:14 Integrated Voltage Measurement... 16:30:23 Voltage Offset Correction: v_off_ini = 1613.976 uV, v_off_fin = 1620.628 uV 16:30:23 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -165.206 uVs 16:30:23 Integrated Voltage Measurement... 16:30:32 Voltage Offset Correction: v_off_ini = 1621.848 uV, v_off_fin = 1628.444 uV 16:30:32 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 162.768 uVs 16:30:32 Integrated Voltage Measurement... 16:30:41 Voltage Offset Correction: v_off_ini = 1631.313 uV, v_off_fin = 1621.793 uV 16:30:41 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -163.607 uVs 16:30:41 Integrated Voltage Measurement... 16:30:50 Voltage Offset Correction: v_off_ini = 1621.023 uV, v_off_fin = 1649.594 uV 16:30:50 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 169.586 uVs 16:30:50 Integrated Voltage Measurement... 16:30:59 Voltage Offset Correction: v_off_ini = 1651.466 uV, v_off_fin = 1641.560 uV 16:30:59 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -157.068 uVs 16:30:59 Integrated Voltage Measurement... 16:31:08 Voltage Offset Correction: v_off_ini = 1646.596 uV, v_off_fin = 1631.837 uV 16:31:08 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 158.521 uVs 16:31:10 I2X Average and RMS of Measurements VT = -162.793 +- 3.691 uVs I2X = 14.977 +- 0.340 uTm^2 16:31:15 Wire Position: X0 = -2.0000 mm, Y0 = 0.0000 mm 16:31:18 Integrated Voltage Measurement... 16:31:27 Voltage Offset Correction: v_off_ini = 1658.393 uV, v_off_fin = 1661.182 uV 16:31:27 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -196.920 uVs 16:31:27 Integrated Voltage Measurement... 16:31:36 Voltage Offset Correction: v_off_ini = 1654.194 uV, v_off_fin = 1660.799 uV 16:31:36 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 183.786 uVs 16:31:36 Integrated Voltage Measurement... 16:31:45 Voltage Offset Correction: v_off_ini = 1652.833 uV, v_off_fin = 1645.918 uV 16:31:45 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -201.342 uVs 16:31:45 Integrated Voltage Measurement... 16:31:54 Voltage Offset Correction: v_off_ini = 1649.390 uV, v_off_fin = 1663.123 uV 16:31:54 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 192.364 uVs 16:31:54 Integrated Voltage Measurement... 16:32:03 Voltage Offset Correction: v_off_ini = 1670.284 uV, v_off_fin = 1658.009 uV 16:32:03 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -194.507 uVs 16:32:03 Integrated Voltage Measurement... 16:32:12 Voltage Offset Correction: v_off_ini = 1653.383 uV, v_off_fin = 1653.494 uV 16:32:12 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 180.665 uVs 16:32:15 I2Y Average and RMS of Measurements VT = -191.597 +- 3.884 uVs I2Y = -17.627 +- 0.357 uTm^2 16:32:19 Wire Position: X0 = -1.0000 mm, Y0 = 0.0000 mm 16:32:23 Integrated Voltage Measurement... 16:32:32 Voltage Offset Correction: v_off_ini = 1675.755 uV, v_off_fin = 1639.197 uV 16:32:32 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -179.959 uVs 16:32:32 Integrated Voltage Measurement... 16:32:41 Voltage Offset Correction: v_off_ini = 1648.466 uV, v_off_fin = 1649.013 uV 16:32:41 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 162.179 uVs 16:32:41 Integrated Voltage Measurement... 16:32:50 Voltage Offset Correction: v_off_ini = 1661.121 uV, v_off_fin = 1669.019 uV 16:32:50 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -173.054 uVs 16:32:50 Integrated Voltage Measurement... 16:32:59 Voltage Offset Correction: v_off_ini = 1665.311 uV, v_off_fin = 1667.516 uV 16:32:59 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 148.990 uVs 16:32:59 Integrated Voltage Measurement... 16:33:08 Voltage Offset Correction: v_off_ini = 1654.887 uV, v_off_fin = 1661.470 uV 16:33:08 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.4999 mm dXY = -0.4999 mm VT = -149.130 uVs 16:33:08 Integrated Voltage Measurement... 16:33:17 Voltage Offset Correction: v_off_ini = 1632.291 uV, v_off_fin = 1635.265 uV 16:33:17 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 148.976 uVs 16:33:22 I1X Average and RMS of Measurements VT = -160.381 +- 8.999 uVs I1X = 32.077 +- 1.799 uTm 16:33:24 Wire Position: X0 = -1.0000 mm, Y0 = 0.0000 mm 16:33:28 Integrated Voltage Measurement... 16:33:37 Voltage Offset Correction: v_off_ini = 1624.782 uV, v_off_fin = 1621.053 uV 16:33:37 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -222.167 uVs 16:33:37 Integrated Voltage Measurement... 16:33:46 Voltage Offset Correction: v_off_ini = 1653.424 uV, v_off_fin = 1661.566 uV 16:33:46 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 215.053 uVs 16:33:46 Integrated Voltage Measurement... 16:33:55 Voltage Offset Correction: v_off_ini = 1656.836 uV, v_off_fin = 1634.454 uV 16:33:55 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -217.166 uVs 16:33:55 Integrated Voltage Measurement... 16:34:04 Voltage Offset Correction: v_off_ini = 1656.485 uV, v_off_fin = 1660.876 uV 16:34:04 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 215.645 uVs 16:34:04 Integrated Voltage Measurement... 16:34:13 Voltage Offset Correction: v_off_ini = 1657.180 uV, v_off_fin = 1648.616 uV 16:34:13 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -215.480 uVs 16:34:13 Integrated Voltage Measurement... 16:34:22 Voltage Offset Correction: v_off_ini = 1636.924 uV, v_off_fin = 1652.661 uV 16:34:22 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 223.239 uVs 16:34:26 I1Y Average and RMS of Measurements VT = -218.125 +- 1.254 uVs I1Y = -43.625 +- 0.251 uTm 16:34:31 Wire Position: X0 = -1.0000 mm, Y0 = 0.0000 mm 16:34:34 Integrated Voltage Measurement... 16:34:43 Voltage Offset Correction: v_off_ini = 1652.030 uV, v_off_fin = 1643.464 uV 16:34:43 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -128.382 uVs 16:34:43 Integrated Voltage Measurement... 16:34:52 Voltage Offset Correction: v_off_ini = 1647.005 uV, v_off_fin = 1662.108 uV 16:34:52 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 153.757 uVs 16:34:52 Integrated Voltage Measurement... 16:35:01 Voltage Offset Correction: v_off_ini = 1637.514 uV, v_off_fin = 1645.501 uV 16:35:01 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -129.327 uVs 16:35:01 Integrated Voltage Measurement... 16:35:10 Voltage Offset Correction: v_off_ini = 1628.170 uV, v_off_fin = 1630.157 uV 16:35:10 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 114.072 uVs 16:35:10 Integrated Voltage Measurement... 16:35:19 Voltage Offset Correction: v_off_ini = 1619.745 uV, v_off_fin = 1625.976 uV 16:35:19 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -136.654 uVs 16:35:19 Integrated Voltage Measurement... 16:35:28 Voltage Offset Correction: v_off_ini = 1625.696 uV, v_off_fin = 1638.514 uV 16:35:28 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 138.589 uVs 16:35:31 I2X Average and RMS of Measurements VT = -133.463 +- 8.437 uVs I2X = 12.279 +- 0.776 uTm^2 16:35:36 Wire Position: X0 = -1.0000 mm, Y0 = 0.0000 mm 16:35:38 Integrated Voltage Measurement... 16:35:47 Voltage Offset Correction: v_off_ini = 1611.726 uV, v_off_fin = 1611.555 uV 16:35:47 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -216.433 uVs 16:35:47 Integrated Voltage Measurement... 16:35:56 Voltage Offset Correction: v_off_ini = 1615.313 uV, v_off_fin = 1606.546 uV 16:35:56 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 216.040 uVs 16:35:56 Integrated Voltage Measurement... 16:36:05 Voltage Offset Correction: v_off_ini = 1590.603 uV, v_off_fin = 1617.417 uV 16:36:05 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -235.212 uVs 16:36:05 Integrated Voltage Measurement... 16:36:14 Voltage Offset Correction: v_off_ini = 1629.144 uV, v_off_fin = 1629.714 uV 16:36:14 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 216.139 uVs 16:36:14 Integrated Voltage Measurement... 16:36:23 Voltage Offset Correction: v_off_ini = 1616.408 uV, v_off_fin = 1627.190 uV 16:36:23 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -192.329 uVs 16:36:23 Integrated Voltage Measurement... 16:36:32 Voltage Offset Correction: v_off_ini = 1615.892 uV, v_off_fin = 1625.680 uV 16:36:32 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 236.187 uVs 16:36:35 I2Y Average and RMS of Measurements VT = -218.723 +- 4.982 uVs I2Y = -20.123 +- 0.458 uTm^2 16:36:39 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:36:44 Integrated Voltage Measurement... 16:36:52 Voltage Offset Correction: v_off_ini = 1635.827 uV, v_off_fin = 1638.645 uV 16:36:52 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -135.240 uVs 16:36:53 Integrated Voltage Measurement... 16:37:01 Voltage Offset Correction: v_off_ini = 1647.247 uV, v_off_fin = 1629.454 uV 16:37:01 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 121.292 uVs 16:37:02 Integrated Voltage Measurement... 16:37:10 Voltage Offset Correction: v_off_ini = 1640.161 uV, v_off_fin = 1632.546 uV 16:37:10 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -117.694 uVs 16:37:11 Integrated Voltage Measurement... 16:37:19 Voltage Offset Correction: v_off_ini = 1627.782 uV, v_off_fin = 1626.939 uV 16:37:19 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 121.389 uVs 16:37:19 Integrated Voltage Measurement... 16:37:28 Voltage Offset Correction: v_off_ini = 1621.587 uV, v_off_fin = 1630.539 uV 16:37:28 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -127.893 uVs 16:37:28 Integrated Voltage Measurement... 16:37:37 Voltage Offset Correction: v_off_ini = 1638.184 uV, v_off_fin = 1647.137 uV 16:37:37 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 128.235 uVs 16:37:42 I1X Average and RMS of Measurements VT = -125.290 +- 4.066 uVs I1X = 25.059 +- 0.814 uTm 16:37:44 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:37:48 Integrated Voltage Measurement... 16:37:57 Voltage Offset Correction: v_off_ini = 1635.662 uV, v_off_fin = 1639.204 uV 16:37:57 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = -231.118 uVs 16:37:57 Integrated Voltage Measurement... 16:38:06 Voltage Offset Correction: v_off_ini = 1636.786 uV, v_off_fin = 1644.548 uV 16:38:06 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 228.713 uVs 16:38:06 Integrated Voltage Measurement... 16:38:15 Voltage Offset Correction: v_off_ini = 1643.151 uV, v_off_fin = 1621.523 uV 16:38:15 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -216.785 uVs 16:38:15 Integrated Voltage Measurement... 16:38:24 Voltage Offset Correction: v_off_ini = 1633.430 uV, v_off_fin = 1633.103 uV 16:38:24 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 208.000 uVs 16:38:24 Integrated Voltage Measurement... 16:38:33 Voltage Offset Correction: v_off_ini = 1653.679 uV, v_off_fin = 1632.620 uV 16:38:33 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -219.771 uVs 16:38:33 Integrated Voltage Measurement... 16:38:42 Voltage Offset Correction: v_off_ini = 1639.200 uV, v_off_fin = 1658.196 uV 16:38:42 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 228.767 uVs 16:38:46 I1Y Average and RMS of Measurements VT = -222.192 +- 7.303 uVs I1Y = -44.439 +- 1.461 uTm 16:38:52 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:38:54 Integrated Voltage Measurement... 16:39:03 Voltage Offset Correction: v_off_ini = 1643.865 uV, v_off_fin = 1649.070 uV 16:39:03 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -102.797 uVs 16:39:03 Integrated Voltage Measurement... 16:39:12 Voltage Offset Correction: v_off_ini = 1649.913 uV, v_off_fin = 1665.104 uV 16:39:12 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 135.115 uVs 16:39:12 Integrated Voltage Measurement... 16:39:21 Voltage Offset Correction: v_off_ini = 1626.621 uV, v_off_fin = 1629.099 uV 16:39:21 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -94.416 uVs 16:39:21 Integrated Voltage Measurement... 16:39:30 Voltage Offset Correction: v_off_ini = 1654.750 uV, v_off_fin = 1659.784 uV 16:39:30 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 113.351 uVs 16:39:30 Integrated Voltage Measurement... 16:39:39 Voltage Offset Correction: v_off_ini = 1668.611 uV, v_off_fin = 1650.246 uV 16:39:39 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -108.232 uVs 16:39:39 Integrated Voltage Measurement... 16:39:48 Voltage Offset Correction: v_off_ini = 1666.021 uV, v_off_fin = 1656.837 uV 16:39:48 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 99.242 uVs 16:39:51 I2X Average and RMS of Measurements VT = -108.859 +- 7.140 uVs I2X = 10.015 +- 0.657 uTm^2 16:39:56 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:39:58 Integrated Voltage Measurement... 16:40:07 Voltage Offset Correction: v_off_ini = 1668.024 uV, v_off_fin = 1647.761 uV 16:40:07 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -223.563 uVs 16:40:07 Integrated Voltage Measurement... 16:40:16 Voltage Offset Correction: v_off_ini = 1668.168 uV, v_off_fin = 1668.643 uV 16:40:16 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 233.939 uVs 16:40:16 Integrated Voltage Measurement... 16:40:25 Voltage Offset Correction: v_off_ini = 1673.724 uV, v_off_fin = 1670.480 uV 16:40:25 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -207.041 uVs 16:40:25 Integrated Voltage Measurement... 16:40:34 Voltage Offset Correction: v_off_ini = 1659.479 uV, v_off_fin = 1654.955 uV 16:40:34 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 215.981 uVs 16:40:34 Integrated Voltage Measurement... 16:40:43 Voltage Offset Correction: v_off_ini = 1662.397 uV, v_off_fin = 1673.342 uV 16:40:43 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -239.007 uVs 16:40:43 Integrated Voltage Measurement... 16:40:52 Voltage Offset Correction: v_off_ini = 1689.741 uV, v_off_fin = 1687.906 uV 16:40:52 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 211.716 uVs 16:40:55 I2Y Average and RMS of Measurements VT = -221.875 +- 7.458 uVs I2Y = -20.412 +- 0.686 uTm^2 16:40:59 Wire Position: X0 = 1.0000 mm, Y0 = 0.0000 mm 16:41:04 Integrated Voltage Measurement... 16:41:13 Voltage Offset Correction: v_off_ini = 1657.165 uV, v_off_fin = 1669.888 uV 16:41:13 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4998 mm dXY = -0.4999 mm VT = -112.042 uVs 16:41:13 Integrated Voltage Measurement... 16:41:22 Voltage Offset Correction: v_off_ini = 1668.440 uV, v_off_fin = 1673.323 uV 16:41:22 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 113.493 uVs 16:41:22 Integrated Voltage Measurement... 16:41:31 Voltage Offset Correction: v_off_ini = 1690.104 uV, v_off_fin = 1684.454 uV 16:41:31 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -92.407 uVs 16:41:31 Integrated Voltage Measurement... 16:41:40 Voltage Offset Correction: v_off_ini = 1677.467 uV, v_off_fin = 1670.272 uV 16:41:40 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 118.783 uVs 16:41:40 Integrated Voltage Measurement... 16:41:49 Voltage Offset Correction: v_off_ini = 1661.578 uV, v_off_fin = 1656.354 uV 16:41:49 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -90.789 uVs 16:41:49 Integrated Voltage Measurement... 16:41:58 Voltage Offset Correction: v_off_ini = 1674.194 uV, v_off_fin = 1661.488 uV 16:41:58 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 109.380 uVs 16:42:02 I1X Average and RMS of Measurements VT = -106.149 +- 5.192 uVs I1X = 21.231 +- 1.039 uTm 16:42:04 Wire Position: X0 = 1.0000 mm, Y0 = 0.0000 mm 16:42:09 Integrated Voltage Measurement... 16:42:17 Voltage Offset Correction: v_off_ini = 1662.460 uV, v_off_fin = 1649.538 uV 16:42:18 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -266.912 uVs 16:42:18 Integrated Voltage Measurement... 16:42:26 Voltage Offset Correction: v_off_ini = 1666.660 uV, v_off_fin = 1675.683 uV 16:42:27 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 230.740 uVs 16:42:27 Integrated Voltage Measurement... 16:42:35 Voltage Offset Correction: v_off_ini = 1660.994 uV, v_off_fin = 1647.059 uV 16:42:35 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -231.641 uVs 16:42:36 Integrated Voltage Measurement... 16:42:44 Voltage Offset Correction: v_off_ini = 1678.469 uV, v_off_fin = 1671.803 uV 16:42:44 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 235.588 uVs 16:42:45 Integrated Voltage Measurement... 16:42:53 Voltage Offset Correction: v_off_ini = 1661.234 uV, v_off_fin = 1658.872 uV 16:42:53 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -236.194 uVs 16:42:53 Integrated Voltage Measurement... 16:43:02 Voltage Offset Correction: v_off_ini = 1666.385 uV, v_off_fin = 1680.882 uV 16:43:02 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 235.881 uVs 16:43:07 I1Y Average and RMS of Measurements VT = -239.493 +- 6.673 uVs I1Y = -47.899 +- 1.336 uTm 16:43:12 Wire Position: X0 = 1.0000 mm, Y0 = 0.0000 mm 16:43:14 Integrated Voltage Measurement... 16:43:23 Voltage Offset Correction: v_off_ini = 1642.331 uV, v_off_fin = 1655.726 uV 16:43:23 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -73.691 uVs 16:43:23 Integrated Voltage Measurement... 16:43:32 Voltage Offset Correction: v_off_ini = 1660.937 uV, v_off_fin = 1677.737 uV 16:43:32 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 80.420 uVs 16:43:32 Integrated Voltage Measurement... 16:43:41 Voltage Offset Correction: v_off_ini = 1674.624 uV, v_off_fin = 1671.440 uV 16:43:41 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -82.758 uVs 16:43:41 Integrated Voltage Measurement... 16:43:50 Voltage Offset Correction: v_off_ini = 1675.299 uV, v_off_fin = 1677.762 uV 16:43:50 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 68.247 uVs 16:43:50 Integrated Voltage Measurement... 16:43:59 Voltage Offset Correction: v_off_ini = 1677.282 uV, v_off_fin = 1662.641 uV 16:43:59 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -82.973 uVs 16:43:59 Integrated Voltage Measurement... 16:44:08 Voltage Offset Correction: v_off_ini = 1667.521 uV, v_off_fin = 1682.337 uV 16:44:08 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 70.946 uVs 16:44:11 I2X Average and RMS of Measurements VT = -76.506 +- 0.710 uVs I2X = 7.039 +- 0.066 uTm^2 16:44:16 Wire Position: X0 = 1.0000 mm, Y0 = 0.0000 mm 16:44:18 Integrated Voltage Measurement... 16:44:27 Voltage Offset Correction: v_off_ini = 1687.726 uV, v_off_fin = 1693.764 uV 16:44:27 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -234.143 uVs 16:44:28 Integrated Voltage Measurement... 16:44:36 Voltage Offset Correction: v_off_ini = 1694.199 uV, v_off_fin = 1671.282 uV 16:44:36 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 235.323 uVs 16:44:37 Integrated Voltage Measurement... 16:44:45 Voltage Offset Correction: v_off_ini = 1674.091 uV, v_off_fin = 1679.915 uV 16:44:46 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -239.875 uVs 16:44:46 Integrated Voltage Measurement... 16:44:55 Voltage Offset Correction: v_off_ini = 1681.560 uV, v_off_fin = 1715.210 uV 16:44:55 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 216.155 uVs 16:44:55 Integrated Voltage Measurement... 16:45:04 Voltage Offset Correction: v_off_ini = 1679.637 uV, v_off_fin = 1666.727 uV 16:45:04 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -230.854 uVs 16:45:04 Integrated Voltage Measurement... 16:45:13 Voltage Offset Correction: v_off_ini = 1677.669 uV, v_off_fin = 1683.663 uV 16:45:13 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 259.397 uVs 16:45:15 I2Y Average and RMS of Measurements VT = -235.958 +- 7.039 uVs I2Y = -21.708 +- 0.648 uTm^2 16:45:20 Wire Position: X0 = 2.0000 mm, Y0 = 0.0000 mm 16:45:24 Integrated Voltage Measurement... 16:45:33 Voltage Offset Correction: v_off_ini = 1722.559 uV, v_off_fin = 1696.784 uV 16:45:33 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -65.666 uVs 16:45:33 Integrated Voltage Measurement... 16:45:42 Voltage Offset Correction: v_off_ini = 1703.779 uV, v_off_fin = 1726.439 uV 16:45:42 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = 107.824 uVs 16:45:42 Integrated Voltage Measurement... 16:45:51 Voltage Offset Correction: v_off_ini = 1693.929 uV, v_off_fin = 1688.360 uV 16:45:51 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -79.587 uVs 16:45:51 Integrated Voltage Measurement... 16:46:00 Voltage Offset Correction: v_off_ini = 1690.090 uV, v_off_fin = 1681.606 uV 16:46:00 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 73.106 uVs 16:46:00 Integrated Voltage Measurement... 16:46:09 Voltage Offset Correction: v_off_ini = 1665.407 uV, v_off_fin = 1688.429 uV 16:46:09 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -97.092 uVs 16:46:09 Integrated Voltage Measurement... 16:46:18 Voltage Offset Correction: v_off_ini = 1688.553 uV, v_off_fin = 1694.953 uV 16:46:18 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 88.134 uVs 16:46:22 I1X Average and RMS of Measurements VT = -85.235 +- 6.726 uVs I1X = 17.048 +- 1.345 uTm 16:46:25 Wire Position: X0 = 2.0000 mm, Y0 = 0.0000 mm 16:46:29 Integrated Voltage Measurement... 16:46:38 Voltage Offset Correction: v_off_ini = 1708.275 uV, v_off_fin = 1699.190 uV 16:46:38 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -254.481 uVs 16:46:38 Integrated Voltage Measurement... 16:46:47 Voltage Offset Correction: v_off_ini = 1696.466 uV, v_off_fin = 1732.726 uV 16:46:47 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 249.592 uVs 16:46:47 Integrated Voltage Measurement... 16:46:56 Voltage Offset Correction: v_off_ini = 1694.794 uV, v_off_fin = 1701.246 uV 16:46:56 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -237.969 uVs 16:46:56 Integrated Voltage Measurement... 16:47:05 Voltage Offset Correction: v_off_ini = 1721.462 uV, v_off_fin = 1708.691 uV 16:47:05 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 247.097 uVs 16:47:05 Integrated Voltage Measurement... 16:47:14 Voltage Offset Correction: v_off_ini = 1718.443 uV, v_off_fin = 1713.086 uV 16:47:14 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -260.956 uVs 16:47:14 Integrated Voltage Measurement... 16:47:23 Voltage Offset Correction: v_off_ini = 1717.475 uV, v_off_fin = 1740.425 uV 16:47:23 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 236.552 uVs 16:47:27 I1Y Average and RMS of Measurements VT = -247.774 +- 3.941 uVs I1Y = -49.555 +- 0.788 uTm 16:47:32 Wire Position: X0 = 2.0000 mm, Y0 = 0.0000 mm 16:47:35 Integrated Voltage Measurement... 16:47:44 Voltage Offset Correction: v_off_ini = 1707.863 uV, v_off_fin = 1724.416 uV 16:47:44 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -77.802 uVs 16:47:44 Integrated Voltage Measurement... 16:47:53 Voltage Offset Correction: v_off_ini = 1723.530 uV, v_off_fin = 1742.214 uV 16:47:53 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 60.698 uVs 16:47:53 Integrated Voltage Measurement... 16:48:02 Voltage Offset Correction: v_off_ini = 1737.358 uV, v_off_fin = 1715.164 uV 16:48:02 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -62.106 uVs 16:48:02 Integrated Voltage Measurement... 16:48:11 Voltage Offset Correction: v_off_ini = 1740.480 uV, v_off_fin = 1744.075 uV 16:48:11 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 59.487 uVs 16:48:11 Integrated Voltage Measurement... 16:48:20 Voltage Offset Correction: v_off_ini = 1734.009 uV, v_off_fin = 1718.542 uV 16:48:20 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -52.486 uVs 16:48:20 Integrated Voltage Measurement... 16:48:29 Voltage Offset Correction: v_off_ini = 1738.331 uV, v_off_fin = 1745.612 uV 16:48:29 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 59.694 uVs 16:48:31 I2X Average and RMS of Measurements VT = -62.045 +- 5.445 uVs I2X = 5.708 +- 0.501 uTm^2 16:48:36 Wire Position: X0 = 2.0000 mm, Y0 = 0.0000 mm 16:48:39 Integrated Voltage Measurement... 16:48:48 Voltage Offset Correction: v_off_ini = 1719.909 uV, v_off_fin = 1733.229 uV 16:48:48 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -245.241 uVs 16:48:48 Integrated Voltage Measurement... 16:48:57 Voltage Offset Correction: v_off_ini = 1720.490 uV, v_off_fin = 1726.853 uV 16:48:57 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 278.316 uVs 16:48:57 Integrated Voltage Measurement... 16:49:06 Voltage Offset Correction: v_off_ini = 1742.669 uV, v_off_fin = 1746.595 uV 16:49:06 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -244.318 uVs 16:49:06 Integrated Voltage Measurement... 16:49:15 Voltage Offset Correction: v_off_ini = 1756.289 uV, v_off_fin = 1743.902 uV 16:49:15 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 233.767 uVs 16:49:15 Integrated Voltage Measurement... 16:49:24 Voltage Offset Correction: v_off_ini = 1744.832 uV, v_off_fin = 1743.135 uV 16:49:24 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -237.393 uVs 16:49:24 Integrated Voltage Measurement... 16:49:33 Voltage Offset Correction: v_off_ini = 1757.467 uV, v_off_fin = 1731.191 uV 16:49:33 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 256.235 uVs 16:49:35 I2Y Average and RMS of Measurements VT = -249.212 +- 9.435 uVs I2Y = -22.927 +- 0.868 uTm^2 16:49:43 Wire Position: X0 = 0.0000 mm, Y0 = -1.0000 mm 16:49:47 Integrated Voltage Measurement... 16:49:56 Voltage Offset Correction: v_off_ini = 1717.471 uV, v_off_fin = 1684.747 uV 16:49:56 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -138.721 uVs 16:49:56 Integrated Voltage Measurement... 16:50:05 Voltage Offset Correction: v_off_ini = 1711.913 uV, v_off_fin = 1711.600 uV 16:50:05 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 139.934 uVs 16:50:05 Integrated Voltage Measurement... 16:50:14 Voltage Offset Correction: v_off_ini = 1729.152 uV, v_off_fin = 1742.265 uV 16:50:14 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -122.659 uVs 16:50:14 Integrated Voltage Measurement... 16:50:23 Voltage Offset Correction: v_off_ini = 1741.723 uV, v_off_fin = 1748.099 uV 16:50:23 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 140.651 uVs 16:50:23 Integrated Voltage Measurement... 16:50:32 Voltage Offset Correction: v_off_ini = 1753.478 uV, v_off_fin = 1742.550 uV 16:50:32 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -128.921 uVs 16:50:32 Integrated Voltage Measurement... 16:50:41 Voltage Offset Correction: v_off_ini = 1754.260 uV, v_off_fin = 1763.053 uV 16:50:41 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 143.037 uVs 16:50:45 I1X Average and RMS of Measurements VT = -135.654 +- 3.141 uVs I1X = 27.131 +- 0.629 uTm 16:50:48 Wire Position: X0 = 0.0000 mm, Y0 = -1.0000 mm 16:50:52 Integrated Voltage Measurement... 16:51:01 Voltage Offset Correction: v_off_ini = 1747.613 uV, v_off_fin = 1743.151 uV 16:51:01 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -265.455 uVs 16:51:01 Integrated Voltage Measurement... 16:51:10 Voltage Offset Correction: v_off_ini = 1741.998 uV, v_off_fin = 1765.358 uV 16:51:10 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 262.996 uVs 16:51:10 Integrated Voltage Measurement... 16:51:19 Voltage Offset Correction: v_off_ini = 1744.056 uV, v_off_fin = 1732.089 uV 16:51:19 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -248.473 uVs 16:51:19 Integrated Voltage Measurement... 16:51:28 Voltage Offset Correction: v_off_ini = 1722.034 uV, v_off_fin = 1721.959 uV 16:51:28 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 254.004 uVs 16:51:28 Integrated Voltage Measurement... 16:51:37 Voltage Offset Correction: v_off_ini = 1741.148 uV, v_off_fin = 1719.653 uV 16:51:37 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -248.665 uVs 16:51:37 Integrated Voltage Measurement... 16:51:46 Voltage Offset Correction: v_off_ini = 1748.328 uV, v_off_fin = 1749.260 uV 16:51:46 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 252.816 uVs 16:51:50 I1Y Average and RMS of Measurements VT = -255.402 +- 6.243 uVs I1Y = -51.080 +- 1.249 uTm 16:51:53 Wire Position: X0 = 0.0000 mm, Y0 = -1.0000 mm 16:51:55 Integrated Voltage Measurement... 16:52:04 Voltage Offset Correction: v_off_ini = 1741.650 uV, v_off_fin = 1733.633 uV 16:52:04 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -121.961 uVs 16:52:04 Integrated Voltage Measurement... 16:52:13 Voltage Offset Correction: v_off_ini = 1745.048 uV, v_off_fin = 1744.608 uV 16:52:13 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 122.118 uVs 16:52:13 Integrated Voltage Measurement... 16:52:22 Voltage Offset Correction: v_off_ini = 1750.654 uV, v_off_fin = 1741.986 uV 16:52:22 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -122.955 uVs 16:52:23 Integrated Voltage Measurement... 16:52:31 Voltage Offset Correction: v_off_ini = 1728.766 uV, v_off_fin = 1736.523 uV 16:52:32 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 100.743 uVs 16:52:32 Integrated Voltage Measurement... 16:52:41 Voltage Offset Correction: v_off_ini = 1721.672 uV, v_off_fin = 1715.253 uV 16:52:41 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -130.877 uVs 16:52:41 Integrated Voltage Measurement... 16:52:50 Voltage Offset Correction: v_off_ini = 1731.504 uV, v_off_fin = 1743.732 uV 16:52:50 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 80.972 uVs 16:52:52 I2X Average and RMS of Measurements VT = -113.271 +- 6.655 uVs I2X = 10.421 +- 0.612 uTm^2 16:52:54 Wire Position: X0 = 0.0000 mm, Y0 = -1.0000 mm 16:52:57 Integrated Voltage Measurement... 16:53:06 Voltage Offset Correction: v_off_ini = 1720.811 uV, v_off_fin = 1734.975 uV 16:53:06 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -263.698 uVs 16:53:06 Integrated Voltage Measurement... 16:53:15 Voltage Offset Correction: v_off_ini = 1732.811 uV, v_off_fin = 1714.613 uV 16:53:15 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 250.089 uVs 16:53:15 Integrated Voltage Measurement... 16:53:24 Voltage Offset Correction: v_off_ini = 1724.380 uV, v_off_fin = 1731.361 uV 16:53:24 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -246.403 uVs 16:53:24 Integrated Voltage Measurement... 16:53:33 Voltage Offset Correction: v_off_ini = 1741.793 uV, v_off_fin = 1738.336 uV 16:53:33 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 254.790 uVs 16:53:33 Integrated Voltage Measurement... 16:53:42 Voltage Offset Correction: v_off_ini = 1742.887 uV, v_off_fin = 1750.499 uV 16:53:42 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -251.815 uVs 16:53:42 Integrated Voltage Measurement... 16:53:51 Voltage Offset Correction: v_off_ini = 1757.743 uV, v_off_fin = 1765.396 uV 16:53:51 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 255.450 uVs 16:53:54 I2Y Average and RMS of Measurements VT = -253.707 +- 2.571 uVs I2Y = -23.341 +- 0.237 uTm^2 16:53:58 Wire Position: X0 = 0.0000 mm, Y0 = -0.5000 mm 16:54:02 Integrated Voltage Measurement... 16:54:11 Voltage Offset Correction: v_off_ini = 1747.857 uV, v_off_fin = 1754.129 uV 16:54:11 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.4998 mm dXY = -0.4999 mm VT = -122.148 uVs 16:54:11 Integrated Voltage Measurement... 16:54:20 Voltage Offset Correction: v_off_ini = 1742.911 uV, v_off_fin = 1755.655 uV 16:54:20 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 127.995 uVs 16:54:20 Integrated Voltage Measurement... 16:54:29 Voltage Offset Correction: v_off_ini = 1743.763 uV, v_off_fin = 1749.094 uV 16:54:29 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -132.709 uVs 16:54:29 Integrated Voltage Measurement... 16:54:38 Voltage Offset Correction: v_off_ini = 1766.908 uV, v_off_fin = 1768.747 uV 16:54:38 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 112.102 uVs 16:54:38 Integrated Voltage Measurement... 16:54:47 Voltage Offset Correction: v_off_ini = 1754.380 uV, v_off_fin = 1766.684 uV 16:54:47 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -116.654 uVs 16:54:47 Integrated Voltage Measurement... 16:54:56 Voltage Offset Correction: v_off_ini = 1756.624 uV, v_off_fin = 1756.444 uV 16:54:56 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 145.202 uVs 16:55:00 I1X Average and RMS of Measurements VT = -126.135 +- 3.560 uVs I1X = 25.229 +- 0.711 uTm 16:55:03 Wire Position: X0 = 0.0000 mm, Y0 = -0.5000 mm 16:55:07 Integrated Voltage Measurement... 16:55:16 Voltage Offset Correction: v_off_ini = 1753.902 uV, v_off_fin = 1740.431 uV 16:55:16 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -236.032 uVs 16:55:16 Integrated Voltage Measurement... 16:55:25 Voltage Offset Correction: v_off_ini = 1739.325 uV, v_off_fin = 1758.352 uV 16:55:25 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 262.217 uVs 16:55:25 Integrated Voltage Measurement... 16:55:34 Voltage Offset Correction: v_off_ini = 1759.808 uV, v_off_fin = 1747.211 uV 16:55:34 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -245.444 uVs 16:55:34 Integrated Voltage Measurement... 16:55:43 Voltage Offset Correction: v_off_ini = 1756.759 uV, v_off_fin = 1766.030 uV 16:55:43 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 239.938 uVs 16:55:43 Integrated Voltage Measurement... 16:55:52 Voltage Offset Correction: v_off_ini = 1750.908 uV, v_off_fin = 1738.629 uV 16:55:52 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -227.603 uVs 16:55:52 Integrated Voltage Measurement... 16:56:01 Voltage Offset Correction: v_off_ini = 1759.552 uV, v_off_fin = 1779.283 uV 16:56:01 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 252.104 uVs 16:56:05 I1Y Average and RMS of Measurements VT = -243.890 +- 3.879 uVs I1Y = -48.778 +- 0.776 uTm 16:56:07 Wire Position: X0 = 0.0000 mm, Y0 = -0.5000 mm 16:56:10 Integrated Voltage Measurement... 16:56:19 Voltage Offset Correction: v_off_ini = 1783.937 uV, v_off_fin = 1796.243 uV 16:56:19 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -90.953 uVs 16:56:19 Integrated Voltage Measurement... 16:56:28 Voltage Offset Correction: v_off_ini = 1782.306 uV, v_off_fin = 1765.734 uV 16:56:28 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 108.344 uVs 16:56:28 Integrated Voltage Measurement... 16:56:37 Voltage Offset Correction: v_off_ini = 1755.701 uV, v_off_fin = 1733.447 uV 16:56:37 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -109.818 uVs 16:56:37 Integrated Voltage Measurement... 16:56:46 Voltage Offset Correction: v_off_ini = 1751.995 uV, v_off_fin = 1767.611 uV 16:56:46 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 100.285 uVs 16:56:46 Integrated Voltage Measurement... 16:56:55 Voltage Offset Correction: v_off_ini = 1725.708 uV, v_off_fin = 1724.476 uV 16:56:55 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -123.339 uVs 16:56:55 Integrated Voltage Measurement... 16:57:04 Voltage Offset Correction: v_off_ini = 1730.391 uV, v_off_fin = 1741.806 uV 16:57:04 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 94.567 uVs 16:57:07 I2X Average and RMS of Measurements VT = -104.551 +- 3.815 uVs I2X = 9.619 +- 0.351 uTm^2 16:57:09 Wire Position: X0 = 0.0000 mm, Y0 = -0.5000 mm 16:57:12 Integrated Voltage Measurement... 16:57:21 Voltage Offset Correction: v_off_ini = 1749.492 uV, v_off_fin = 1746.934 uV 16:57:21 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -239.606 uVs 16:57:21 Integrated Voltage Measurement... 16:57:30 Voltage Offset Correction: v_off_ini = 1750.273 uV, v_off_fin = 1763.610 uV 16:57:30 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 226.746 uVs 16:57:30 Integrated Voltage Measurement... 16:57:39 Voltage Offset Correction: v_off_ini = 1734.631 uV, v_off_fin = 1745.797 uV 16:57:39 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -209.205 uVs 16:57:39 Integrated Voltage Measurement... 16:57:48 Voltage Offset Correction: v_off_ini = 1748.497 uV, v_off_fin = 1731.449 uV 16:57:48 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 235.215 uVs 16:57:48 Integrated Voltage Measurement... 16:57:57 Voltage Offset Correction: v_off_ini = 1723.951 uV, v_off_fin = 1722.643 uV 16:57:57 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -246.320 uVs 16:57:57 Integrated Voltage Measurement... 16:58:06 Voltage Offset Correction: v_off_ini = 1719.749 uV, v_off_fin = 1726.013 uV 16:58:06 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 231.267 uVs 16:58:08 I2Y Average and RMS of Measurements VT = -231.393 +- 6.886 uVs I2Y = -21.288 +- 0.634 uTm^2 16:58:12 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:58:16 Integrated Voltage Measurement... 16:58:25 Voltage Offset Correction: v_off_ini = 1717.038 uV, v_off_fin = 1717.245 uV 16:58:25 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -118.636 uVs 16:58:25 Integrated Voltage Measurement... 16:58:34 Voltage Offset Correction: v_off_ini = 1709.031 uV, v_off_fin = 1738.787 uV 16:58:34 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = 133.348 uVs 16:58:34 Integrated Voltage Measurement... 16:58:43 Voltage Offset Correction: v_off_ini = 1707.570 uV, v_off_fin = 1717.974 uV 16:58:43 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -111.478 uVs 16:58:43 Integrated Voltage Measurement... 16:58:52 Voltage Offset Correction: v_off_ini = 1700.586 uV, v_off_fin = 1713.869 uV 16:58:52 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 121.117 uVs 16:58:52 Integrated Voltage Measurement... 16:59:01 Voltage Offset Correction: v_off_ini = 1716.280 uV, v_off_fin = 1722.960 uV 16:59:01 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -148.141 uVs 16:59:01 Integrated Voltage Measurement... 16:59:10 Voltage Offset Correction: v_off_ini = 1717.369 uV, v_off_fin = 1722.574 uV 16:59:10 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 136.934 uVs 16:59:15 I1X Average and RMS of Measurements VT = -128.276 +- 10.833 uVs I1X = 25.656 +- 2.167 uTm 16:59:17 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:59:21 Integrated Voltage Measurement... 16:59:30 Voltage Offset Correction: v_off_ini = 1721.919 uV, v_off_fin = 1706.567 uV 16:59:30 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -221.380 uVs 16:59:30 Integrated Voltage Measurement... 16:59:39 Voltage Offset Correction: v_off_ini = 1707.298 uV, v_off_fin = 1719.646 uV 16:59:39 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 244.187 uVs 16:59:39 Integrated Voltage Measurement... 16:59:48 Voltage Offset Correction: v_off_ini = 1683.633 uV, v_off_fin = 1692.002 uV 16:59:48 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -240.567 uVs 16:59:48 Integrated Voltage Measurement... 16:59:57 Voltage Offset Correction: v_off_ini = 1695.471 uV, v_off_fin = 1725.416 uV 16:59:57 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 216.442 uVs 16:59:57 Integrated Voltage Measurement... 17:00:06 Voltage Offset Correction: v_off_ini = 1711.554 uV, v_off_fin = 1710.584 uV 17:00:06 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -239.120 uVs 17:00:06 Integrated Voltage Measurement... 17:00:15 Voltage Offset Correction: v_off_ini = 1688.123 uV, v_off_fin = 1680.759 uV 17:00:15 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 242.105 uVs 17:00:20 I1Y Average and RMS of Measurements VT = -233.967 +- 5.013 uVs I1Y = -46.792 +- 1.001 uTm 17:00:22 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 17:00:25 Integrated Voltage Measurement... 17:00:34 Voltage Offset Correction: v_off_ini = 1679.441 uV, v_off_fin = 1676.237 uV 17:00:34 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -90.983 uVs 17:00:34 Integrated Voltage Measurement... 17:00:43 Voltage Offset Correction: v_off_ini = 1692.141 uV, v_off_fin = 1696.131 uV 17:00:43 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 92.536 uVs 17:00:43 Integrated Voltage Measurement... 17:00:52 Voltage Offset Correction: v_off_ini = 1695.004 uV, v_off_fin = 1695.595 uV 17:00:52 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -118.874 uVs 17:00:52 Integrated Voltage Measurement... 17:01:01 Voltage Offset Correction: v_off_ini = 1700.842 uV, v_off_fin = 1717.725 uV 17:01:01 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 92.756 uVs 17:01:01 Integrated Voltage Measurement... 17:01:10 Voltage Offset Correction: v_off_ini = 1684.557 uV, v_off_fin = 1681.108 uV 17:01:10 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -96.289 uVs 17:01:10 Integrated Voltage Measurement... 17:01:19 Voltage Offset Correction: v_off_ini = 1672.613 uV, v_off_fin = 1704.580 uV 17:01:19 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 99.578 uVs 17:01:21 I2X Average and RMS of Measurements VT = -98.503 +- 5.752 uVs I2X = 9.062 +- 0.529 uTm^2 17:01:24 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 17:01:26 Integrated Voltage Measurement... 17:01:35 Voltage Offset Correction: v_off_ini = 1693.351 uV, v_off_fin = 1691.682 uV 17:01:35 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -247.980 uVs 17:01:35 Integrated Voltage Measurement... 17:01:44 Voltage Offset Correction: v_off_ini = 1694.210 uV, v_off_fin = 1681.266 uV 17:01:44 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 216.361 uVs 17:01:44 Integrated Voltage Measurement... 17:01:53 Voltage Offset Correction: v_off_ini = 1672.606 uV, v_off_fin = 1695.568 uV 17:01:53 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -225.470 uVs 17:01:53 Integrated Voltage Measurement... 17:02:02 Voltage Offset Correction: v_off_ini = 1686.263 uV, v_off_fin = 1677.985 uV 17:02:02 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 214.915 uVs 17:02:02 Integrated Voltage Measurement... 17:02:11 Voltage Offset Correction: v_off_ini = 1683.974 uV, v_off_fin = 1686.765 uV 17:02:11 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -214.747 uVs 17:02:11 Integrated Voltage Measurement... 17:02:20 Voltage Offset Correction: v_off_ini = 1681.921 uV, v_off_fin = 1705.347 uV 17:02:20 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 233.796 uVs 17:02:23 I2Y Average and RMS of Measurements VT = -225.545 +- 4.972 uVs I2Y = -20.750 +- 0.457 uTm^2 17:02:27 Wire Position: X0 = 0.0000 mm, Y0 = 0.5000 mm 17:02:31 Integrated Voltage Measurement... 17:02:40 Voltage Offset Correction: v_off_ini = 1684.994 uV, v_off_fin = 1682.357 uV 17:02:40 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -117.419 uVs 17:02:40 Integrated Voltage Measurement... 17:02:49 Voltage Offset Correction: v_off_ini = 1684.436 uV, v_off_fin = 1669.633 uV 17:02:49 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 128.410 uVs 17:02:49 Integrated Voltage Measurement... 17:02:58 Voltage Offset Correction: v_off_ini = 1671.145 uV, v_off_fin = 1675.518 uV 17:02:58 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -127.447 uVs 17:02:58 Integrated Voltage Measurement... 17:03:07 Voltage Offset Correction: v_off_ini = 1670.407 uV, v_off_fin = 1688.012 uV 17:03:07 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 133.974 uVs 17:03:07 Integrated Voltage Measurement... 17:03:16 Voltage Offset Correction: v_off_ini = 1679.204 uV, v_off_fin = 1683.741 uV 17:03:16 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -119.694 uVs 17:03:16 Integrated Voltage Measurement... 17:03:25 Voltage Offset Correction: v_off_ini = 1714.052 uV, v_off_fin = 1699.515 uV 17:03:25 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 103.825 uVs 17:03:29 I1X Average and RMS of Measurements VT = -121.795 +- 7.777 uVs I1X = 24.359 +- 1.555 uTm 17:03:32 Wire Position: X0 = 0.0000 mm, Y0 = 0.5000 mm 17:03:36 Integrated Voltage Measurement... 17:03:45 Voltage Offset Correction: v_off_ini = 1704.701 uV, v_off_fin = 1678.852 uV 17:03:45 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -200.057 uVs 17:03:45 Integrated Voltage Measurement... 17:03:54 Voltage Offset Correction: v_off_ini = 1681.973 uV, v_off_fin = 1693.687 uV 17:03:54 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 226.699 uVs 17:03:54 Integrated Voltage Measurement... 17:04:03 Voltage Offset Correction: v_off_ini = 1673.057 uV, v_off_fin = 1671.655 uV 17:04:03 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -198.686 uVs 17:04:03 Integrated Voltage Measurement... 17:04:12 Voltage Offset Correction: v_off_ini = 1693.196 uV, v_off_fin = 1693.043 uV 17:04:12 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 229.534 uVs 17:04:12 Integrated Voltage Measurement... 17:04:21 Voltage Offset Correction: v_off_ini = 1680.574 uV, v_off_fin = 1672.383 uV 17:04:21 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -207.289 uVs 17:04:21 Integrated Voltage Measurement... 17:04:30 Voltage Offset Correction: v_off_ini = 1646.195 uV, v_off_fin = 1665.274 uV 17:04:30 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 211.938 uVs 17:04:34 I1Y Average and RMS of Measurements VT = -212.367 +- 1.970 uVs I1Y = -42.471 +- 0.394 uTm 17:04:37 Wire Position: X0 = 0.0000 mm, Y0 = 0.5000 mm 17:04:39 Integrated Voltage Measurement... 17:04:48 Voltage Offset Correction: v_off_ini = 1691.244 uV, v_off_fin = 1671.131 uV 17:04:48 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -110.284 uVs 17:04:48 Integrated Voltage Measurement... 17:04:57 Voltage Offset Correction: v_off_ini = 1683.340 uV, v_off_fin = 1683.630 uV 17:04:57 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 139.786 uVs 17:04:57 Integrated Voltage Measurement... 17:05:06 Voltage Offset Correction: v_off_ini = 1685.084 uV, v_off_fin = 1685.131 uV 17:05:06 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -96.877 uVs 17:05:06 Integrated Voltage Measurement... 17:05:15 Voltage Offset Correction: v_off_ini = 1718.702 uV, v_off_fin = 1707.457 uV 17:05:15 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 87.604 uVs 17:05:15 Integrated Voltage Measurement... 17:05:24 Voltage Offset Correction: v_off_ini = 1711.626 uV, v_off_fin = 1695.450 uV 17:05:24 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -119.268 uVs 17:05:24 Integrated Voltage Measurement... 17:05:33 Voltage Offset Correction: v_off_ini = 1683.790 uV, v_off_fin = 1693.905 uV 17:05:33 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 105.554 uVs 17:05:36 I2X Average and RMS of Measurements VT = -109.896 +- 13.506 uVs I2X = 10.111 +- 1.243 uTm^2 17:05:38 Wire Position: X0 = 0.0000 mm, Y0 = 0.5000 mm 17:05:41 Integrated Voltage Measurement... 17:05:50 Voltage Offset Correction: v_off_ini = 1697.094 uV, v_off_fin = 1692.088 uV 17:05:50 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -180.874 uVs 17:05:50 Integrated Voltage Measurement... 17:05:59 Voltage Offset Correction: v_off_ini = 1706.017 uV, v_off_fin = 1675.454 uV 17:05:59 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 210.303 uVs 17:05:59 Integrated Voltage Measurement... 17:06:08 Voltage Offset Correction: v_off_ini = 1693.894 uV, v_off_fin = 1706.730 uV 17:06:08 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = -207.762 uVs 17:06:08 Integrated Voltage Measurement... 17:06:17 Voltage Offset Correction: v_off_ini = 1710.015 uV, v_off_fin = 1704.244 uV 17:06:17 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 203.262 uVs 17:06:17 Integrated Voltage Measurement... 17:06:26 Voltage Offset Correction: v_off_ini = 1712.441 uV, v_off_fin = 1717.415 uV 17:06:26 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -207.322 uVs 17:06:26 Integrated Voltage Measurement... 17:06:35 Voltage Offset Correction: v_off_ini = 1722.775 uV, v_off_fin = 1738.486 uV 17:06:35 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 201.132 uVs 17:06:37 I2Y Average and RMS of Measurements VT = -201.776 +- 4.407 uVs I2Y = -18.564 +- 0.406 uTm^2 17:06:41 Wire Position: X0 = 0.0000 mm, Y0 = 1.0000 mm 17:06:46 Integrated Voltage Measurement... 17:06:55 Voltage Offset Correction: v_off_ini = 1719.442 uV, v_off_fin = 1703.621 uV 17:06:55 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -126.662 uVs 17:06:55 Integrated Voltage Measurement... 17:07:04 Voltage Offset Correction: v_off_ini = 1708.126 uV, v_off_fin = 1723.544 uV 17:07:04 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 104.336 uVs 17:07:04 Integrated Voltage Measurement... 17:07:13 Voltage Offset Correction: v_off_ini = 1717.590 uV, v_off_fin = 1725.090 uV 17:07:13 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -100.592 uVs 17:07:13 Integrated Voltage Measurement... 17:07:22 Voltage Offset Correction: v_off_ini = 1719.278 uV, v_off_fin = 1724.666 uV 17:07:22 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 111.803 uVs 17:07:22 Integrated Voltage Measurement... 17:07:31 Voltage Offset Correction: v_off_ini = 1740.555 uV, v_off_fin = 1739.382 uV 17:07:31 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -106.341 uVs 17:07:31 Integrated Voltage Measurement... 17:07:40 Voltage Offset Correction: v_off_ini = 1750.711 uV, v_off_fin = 1734.212 uV 17:07:40 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 104.071 uVs 17:07:44 I1X Average and RMS of Measurements VT = -108.967 +- 4.636 uVs I1X = 21.794 +- 0.928 uTm 17:07:46 Wire Position: X0 = 0.0000 mm, Y0 = 1.0000 mm 17:07:51 Integrated Voltage Measurement... 17:08:00 Voltage Offset Correction: v_off_ini = 1739.058 uV, v_off_fin = 1717.615 uV 17:08:00 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -193.695 uVs 17:08:00 Integrated Voltage Measurement... 17:08:09 Voltage Offset Correction: v_off_ini = 1721.618 uV, v_off_fin = 1725.215 uV 17:08:09 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 197.821 uVs 17:08:09 Integrated Voltage Measurement... 17:08:18 Voltage Offset Correction: v_off_ini = 1717.226 uV, v_off_fin = 1713.949 uV 17:08:18 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -194.306 uVs 17:08:18 Integrated Voltage Measurement... 17:08:27 Voltage Offset Correction: v_off_ini = 1730.907 uV, v_off_fin = 1749.507 uV 17:08:27 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 199.351 uVs 17:08:27 Integrated Voltage Measurement... 17:08:36 Voltage Offset Correction: v_off_ini = 1741.399 uV, v_off_fin = 1725.711 uV 17:08:36 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -224.468 uVs 17:08:36 Integrated Voltage Measurement... 17:08:45 Voltage Offset Correction: v_off_ini = 1706.062 uV, v_off_fin = 1730.017 uV 17:08:45 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 232.051 uVs 17:08:49 I1Y Average and RMS of Measurements VT = -206.949 +- 15.075 uVs I1Y = -41.386 +- 3.015 uTm 17:08:51 Wire Position: X0 = 0.0000 mm, Y0 = 1.0000 mm 17:08:54 Integrated Voltage Measurement... 17:09:03 Voltage Offset Correction: v_off_ini = 1728.379 uV, v_off_fin = 1737.161 uV 17:09:03 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -96.094 uVs 17:09:03 Integrated Voltage Measurement... 17:09:12 Voltage Offset Correction: v_off_ini = 1758.012 uV, v_off_fin = 1756.266 uV 17:09:12 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 97.590 uVs 17:09:12 Integrated Voltage Measurement... 17:09:21 Voltage Offset Correction: v_off_ini = 1742.840 uV, v_off_fin = 1726.801 uV 17:09:21 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -89.215 uVs 17:09:21 Integrated Voltage Measurement... 17:09:30 Voltage Offset Correction: v_off_ini = 1718.191 uV, v_off_fin = 1712.017 uV 17:09:30 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 87.479 uVs 17:09:30 Integrated Voltage Measurement... 17:09:39 Voltage Offset Correction: v_off_ini = 1719.992 uV, v_off_fin = 1698.767 uV 17:09:39 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -86.245 uVs 17:09:39 Integrated Voltage Measurement... 17:09:48 Voltage Offset Correction: v_off_ini = 1722.500 uV, v_off_fin = 1718.381 uV 17:09:48 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 106.378 uVs 17:09:50 I2X Average and RMS of Measurements VT = -93.834 +- 3.886 uVs I2X = 8.633 +- 0.357 uTm^2 17:09:53 Wire Position: X0 = 0.0000 mm, Y0 = 1.0000 mm 17:09:56 Integrated Voltage Measurement... 17:10:04 Voltage Offset Correction: v_off_ini = 1703.912 uV, v_off_fin = 1727.576 uV 17:10:05 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -171.401 uVs 17:10:05 Integrated Voltage Measurement... 17:10:14 Voltage Offset Correction: v_off_ini = 1708.325 uV, v_off_fin = 1733.818 uV 17:10:14 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 210.157 uVs 17:10:14 Integrated Voltage Measurement... 17:10:23 Voltage Offset Correction: v_off_ini = 1724.228 uV, v_off_fin = 1718.096 uV 17:10:23 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -199.996 uVs 17:10:23 Integrated Voltage Measurement... 17:10:32 Voltage Offset Correction: v_off_ini = 1740.849 uV, v_off_fin = 1736.382 uV 17:10:32 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 221.168 uVs 17:10:32 Integrated Voltage Measurement... 17:10:41 Voltage Offset Correction: v_off_ini = 1737.456 uV, v_off_fin = 1739.986 uV 17:10:41 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -185.857 uVs 17:10:41 Integrated Voltage Measurement... 17:10:50 Voltage Offset Correction: v_off_ini = 1717.680 uV, v_off_fin = 1733.870 uV 17:10:50 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 233.613 uVs 17:10:52 I2Y Average and RMS of Measurements VT = -203.699 +- 9.142 uVs I2Y = -18.740 +- 0.841 uTm^2 17:11:08 Wire Position: X0 = -2.0000 mm, Y0 = 0.0000 mm 17:11:12 Integrated Voltage Measurement... 17:11:21 Voltage Offset Correction: v_off_ini = 1716.682 uV, v_off_fin = 1731.528 uV 17:11:21 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -192.339 uVs 17:11:21 Integrated Voltage Measurement... 17:11:30 Voltage Offset Correction: v_off_ini = 1732.479 uV, v_off_fin = 1721.752 uV 17:11:30 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 177.900 uVs 17:11:30 Integrated Voltage Measurement... 17:11:39 Voltage Offset Correction: v_off_ini = 1720.111 uV, v_off_fin = 1744.025 uV 17:11:39 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -166.299 uVs 17:11:39 Integrated Voltage Measurement... 17:11:48 Voltage Offset Correction: v_off_ini = 1753.369 uV, v_off_fin = 1747.780 uV 17:11:48 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 192.035 uVs 17:11:48 Integrated Voltage Measurement... 17:11:57 Voltage Offset Correction: v_off_ini = 1764.394 uV, v_off_fin = 1747.423 uV 17:11:57 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -165.503 uVs 17:11:57 Integrated Voltage Measurement... 17:12:06 Voltage Offset Correction: v_off_ini = 1761.027 uV, v_off_fin = 1782.248 uV 17:12:06 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 176.409 uVs 17:12:10 I1X Average and RMS of Measurements VT = -178.414 +- 5.807 uVs I1X = 35.683 +- 1.162 uTm 17:12:13 Wire Position: X0 = -2.0000 mm, Y0 = 0.0000 mm 17:12:17 Integrated Voltage Measurement... 17:12:26 Voltage Offset Correction: v_off_ini = 1776.807 uV, v_off_fin = 1763.657 uV 17:12:26 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.4999 mm VT = -200.759 uVs 17:12:26 Integrated Voltage Measurement... 17:12:35 Voltage Offset Correction: v_off_ini = 1773.293 uV, v_off_fin = 1761.964 uV 17:12:35 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 183.116 uVs 17:12:35 Integrated Voltage Measurement... 17:12:44 Voltage Offset Correction: v_off_ini = 1749.317 uV, v_off_fin = 1753.407 uV 17:12:44 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -212.174 uVs 17:12:44 Integrated Voltage Measurement... 17:12:53 Voltage Offset Correction: v_off_ini = 1759.561 uV, v_off_fin = 1767.697 uV 17:12:53 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 200.530 uVs 17:12:53 Integrated Voltage Measurement... 17:13:02 Voltage Offset Correction: v_off_ini = 1750.773 uV, v_off_fin = 1739.434 uV 17:13:02 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -204.076 uVs 17:13:02 Integrated Voltage Measurement... 17:13:11 Voltage Offset Correction: v_off_ini = 1759.107 uV, v_off_fin = 1767.956 uV 17:13:11 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 202.513 uVs 17:13:15 I1Y Average and RMS of Measurements VT = -200.528 +- 6.201 uVs I1Y = -40.106 +- 1.239 uTm 17:13:20 Wire Position: X0 = -2.0000 mm, Y0 = 0.0000 mm 17:13:23 Integrated Voltage Measurement... 17:13:32 Voltage Offset Correction: v_off_ini = 1758.983 uV, v_off_fin = 1744.091 uV 17:13:32 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -152.640 uVs 17:13:32 Integrated Voltage Measurement... 17:13:41 Voltage Offset Correction: v_off_ini = 1753.636 uV, v_off_fin = 1773.633 uV 17:13:41 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 153.049 uVs 17:13:41 Integrated Voltage Measurement... 17:13:50 Voltage Offset Correction: v_off_ini = 1767.193 uV, v_off_fin = 1752.773 uV 17:13:50 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -174.661 uVs 17:13:50 Integrated Voltage Measurement... 17:13:59 Voltage Offset Correction: v_off_ini = 1771.666 uV, v_off_fin = 1756.368 uV 17:13:59 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 158.155 uVs 17:13:59 Integrated Voltage Measurement... 17:14:08 Voltage Offset Correction: v_off_ini = 1751.237 uV, v_off_fin = 1737.912 uV 17:14:08 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -156.814 uVs 17:14:08 Integrated Voltage Measurement... 17:14:17 Voltage Offset Correction: v_off_ini = 1769.851 uV, v_off_fin = 1773.518 uV 17:14:17 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 131.182 uVs 17:14:19 I2X Average and RMS of Measurements VT = -154.417 +- 9.216 uVs I2X = 14.206 +- 0.848 uTm^2 17:14:25 Wire Position: X0 = -2.0000 mm, Y0 = 0.0000 mm 17:14:27 Integrated Voltage Measurement... 17:14:36 Voltage Offset Correction: v_off_ini = 1747.375 uV, v_off_fin = 1748.346 uV 17:14:36 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -217.973 uVs 17:14:36 Integrated Voltage Measurement... 17:14:45 Voltage Offset Correction: v_off_ini = 1752.619 uV, v_off_fin = 1761.061 uV 17:14:45 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 190.017 uVs 17:14:45 Integrated Voltage Measurement... 17:14:54 Voltage Offset Correction: v_off_ini = 1763.176 uV, v_off_fin = 1784.201 uV 17:14:54 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -227.829 uVs 17:14:54 Integrated Voltage Measurement... 17:15:03 Voltage Offset Correction: v_off_ini = 1758.554 uV, v_off_fin = 1781.679 uV 17:15:03 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 211.900 uVs 17:15:03 Integrated Voltage Measurement... 17:15:12 Voltage Offset Correction: v_off_ini = 1765.850 uV, v_off_fin = 1765.140 uV 17:15:12 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -192.246 uVs 17:15:12 Integrated Voltage Measurement... 17:15:21 Voltage Offset Correction: v_off_ini = 1761.085 uV, v_off_fin = 1789.921 uV 17:15:21 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 176.929 uVs 17:15:24 I2Y Average and RMS of Measurements VT = -202.816 +- 14.426 uVs I2Y = -18.659 +- 1.327 uTm^2 17:15:28 Wire Position: X0 = -1.0000 mm, Y0 = 0.0000 mm 17:15:33 Integrated Voltage Measurement... 17:15:41 Voltage Offset Correction: v_off_ini = 1760.336 uV, v_off_fin = 1767.578 uV 17:15:42 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.4999 mm dXY = -0.4999 mm VT = -144.862 uVs 17:15:42 Integrated Voltage Measurement... 17:15:50 Voltage Offset Correction: v_off_ini = 1760.531 uV, v_off_fin = 1768.937 uV 17:15:51 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = 136.441 uVs 17:15:51 Integrated Voltage Measurement... 17:15:59 Voltage Offset Correction: v_off_ini = 1769.218 uV, v_off_fin = 1755.251 uV 17:16:00 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -125.906 uVs 17:16:00 Integrated Voltage Measurement... 17:16:08 Voltage Offset Correction: v_off_ini = 1759.358 uV, v_off_fin = 1788.850 uV 17:16:09 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 169.434 uVs 17:16:09 Integrated Voltage Measurement... 17:16:18 Voltage Offset Correction: v_off_ini = 1774.450 uV, v_off_fin = 1772.069 uV 17:16:18 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -145.623 uVs 17:16:18 Integrated Voltage Measurement... 17:16:27 Voltage Offset Correction: v_off_ini = 1758.356 uV, v_off_fin = 1744.585 uV 17:16:27 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 167.005 uVs 17:16:31 I1X Average and RMS of Measurements VT = -148.212 +- 6.406 uVs I1X = 29.644 +- 1.279 uTm 17:16:33 Wire Position: X0 = -1.0000 mm, Y0 = 0.0000 mm 17:16:37 Integrated Voltage Measurement... 17:16:46 Voltage Offset Correction: v_off_ini = 1773.729 uV, v_off_fin = 1779.625 uV 17:16:46 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -201.284 uVs 17:16:46 Integrated Voltage Measurement... 17:16:55 Voltage Offset Correction: v_off_ini = 1790.243 uV, v_off_fin = 1793.139 uV 17:16:55 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 211.619 uVs 17:16:55 Integrated Voltage Measurement... 17:17:04 Voltage Offset Correction: v_off_ini = 1794.500 uV, v_off_fin = 1784.720 uV 17:17:04 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -208.573 uVs 17:17:04 Integrated Voltage Measurement... 17:17:13 Voltage Offset Correction: v_off_ini = 1792.305 uV, v_off_fin = 1794.599 uV 17:17:13 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 199.519 uVs 17:17:13 Integrated Voltage Measurement... 17:17:22 Voltage Offset Correction: v_off_ini = 1801.918 uV, v_off_fin = 1792.196 uV 17:17:22 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -194.763 uVs 17:17:22 Integrated Voltage Measurement... 17:17:31 Voltage Offset Correction: v_off_ini = 1807.185 uV, v_off_fin = 1795.369 uV 17:17:31 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 214.038 uVs 17:17:36 I1Y Average and RMS of Measurements VT = -204.966 +- 1.060 uVs I1Y = -40.993 +- 0.212 uTm 17:17:41 Wire Position: X0 = -1.0000 mm, Y0 = 0.0000 mm 17:17:43 Integrated Voltage Measurement... 17:17:52 Voltage Offset Correction: v_off_ini = 1804.743 uV, v_off_fin = 1799.859 uV 17:17:52 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -119.622 uVs 17:17:52 Integrated Voltage Measurement... 17:18:01 Voltage Offset Correction: v_off_ini = 1820.108 uV, v_off_fin = 1848.139 uV 17:18:01 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 141.957 uVs 17:18:01 Integrated Voltage Measurement... 17:18:10 Voltage Offset Correction: v_off_ini = 1812.635 uV, v_off_fin = 1818.011 uV 17:18:10 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -160.728 uVs 17:18:10 Integrated Voltage Measurement... 17:18:19 Voltage Offset Correction: v_off_ini = 1818.956 uV, v_off_fin = 1850.056 uV 17:18:19 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 130.364 uVs 17:18:19 Integrated Voltage Measurement... 17:18:28 Voltage Offset Correction: v_off_ini = 1837.269 uV, v_off_fin = 1827.085 uV 17:18:28 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -106.207 uVs 17:18:28 Integrated Voltage Measurement... 17:18:37 Voltage Offset Correction: v_off_ini = 1841.516 uV, v_off_fin = 1854.677 uV 17:18:37 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 118.095 uVs 17:18:40 I2X Average and RMS of Measurements VT = -129.495 +- 13.664 uVs I2X = 11.914 +- 1.257 uTm^2 17:18:45 Wire Position: X0 = -1.0000 mm, Y0 = 0.0000 mm 17:18:47 Integrated Voltage Measurement... 17:18:56 Voltage Offset Correction: v_off_ini = 1839.509 uV, v_off_fin = 1845.009 uV 17:18:56 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -200.438 uVs 17:18:56 Integrated Voltage Measurement... 17:19:05 Voltage Offset Correction: v_off_ini = 1840.198 uV, v_off_fin = 1839.334 uV 17:19:05 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 205.321 uVs 17:19:06 Integrated Voltage Measurement... 17:19:14 Voltage Offset Correction: v_off_ini = 1835.005 uV, v_off_fin = 1837.410 uV 17:19:15 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -214.938 uVs 17:19:15 Integrated Voltage Measurement... 17:19:24 Voltage Offset Correction: v_off_ini = 1860.064 uV, v_off_fin = 1840.721 uV 17:19:24 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 215.998 uVs 17:19:24 Integrated Voltage Measurement... 17:19:33 Voltage Offset Correction: v_off_ini = 1867.452 uV, v_off_fin = 1860.903 uV 17:19:33 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -210.961 uVs 17:19:33 Integrated Voltage Measurement... 17:19:42 Voltage Offset Correction: v_off_ini = 1875.860 uV, v_off_fin = 1860.079 uV 17:19:42 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 220.312 uVs 17:19:44 I2Y Average and RMS of Measurements VT = -211.328 +- 5.974 uVs I2Y = -19.442 +- 0.550 uTm^2 17:19:49 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 17:19:53 Integrated Voltage Measurement... 17:20:02 Voltage Offset Correction: v_off_ini = 1851.613 uV, v_off_fin = 1855.880 uV 17:20:02 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -148.784 uVs 17:20:02 Integrated Voltage Measurement... 17:20:11 Voltage Offset Correction: v_off_ini = 1849.238 uV, v_off_fin = 1857.942 uV 17:20:11 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 105.316 uVs 17:20:11 Integrated Voltage Measurement... 17:20:20 Voltage Offset Correction: v_off_ini = 1853.276 uV, v_off_fin = 1854.206 uV 17:20:20 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -121.603 uVs 17:20:20 Integrated Voltage Measurement... 17:20:29 Voltage Offset Correction: v_off_ini = 1865.496 uV, v_off_fin = 1848.517 uV 17:20:29 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 86.524 uVs 17:20:29 Integrated Voltage Measurement... 17:20:38 Voltage Offset Correction: v_off_ini = 1816.886 uV, v_off_fin = 1835.242 uV 17:20:38 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -123.718 uVs 17:20:38 Integrated Voltage Measurement... 17:20:47 Voltage Offset Correction: v_off_ini = 1857.761 uV, v_off_fin = 1846.627 uV 17:20:47 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 115.047 uVs 17:20:51 I1X Average and RMS of Measurements VT = -116.832 +- 9.556 uVs I1X = 23.366 +- 1.911 uTm 17:20:54 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 17:20:58 Integrated Voltage Measurement... 17:21:07 Voltage Offset Correction: v_off_ini = 1839.136 uV, v_off_fin = 1832.642 uV 17:21:07 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -223.012 uVs 17:21:07 Integrated Voltage Measurement... 17:21:16 Voltage Offset Correction: v_off_ini = 1823.947 uV, v_off_fin = 1815.454 uV 17:21:16 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 220.284 uVs 17:21:16 Integrated Voltage Measurement... 17:21:25 Voltage Offset Correction: v_off_ini = 1808.864 uV, v_off_fin = 1806.428 uV 17:21:25 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -223.390 uVs 17:21:25 Integrated Voltage Measurement... 17:21:34 Voltage Offset Correction: v_off_ini = 1814.414 uV, v_off_fin = 1811.161 uV 17:21:34 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 207.398 uVs 17:21:34 Integrated Voltage Measurement... 17:21:43 Voltage Offset Correction: v_off_ini = 1841.807 uV, v_off_fin = 1823.548 uV 17:21:43 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -220.346 uVs 17:21:43 Integrated Voltage Measurement... 17:21:52 Voltage Offset Correction: v_off_ini = 1853.861 uV, v_off_fin = 1863.293 uV 17:21:52 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 217.832 uVs 17:21:56 I1Y Average and RMS of Measurements VT = -218.710 +- 2.567 uVs I1Y = -43.742 +- 0.513 uTm 17:22:01 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 17:22:04 Integrated Voltage Measurement... 17:22:13 Voltage Offset Correction: v_off_ini = 1820.899 uV, v_off_fin = 1809.559 uV 17:22:13 Integrated Voltage Measurement dXYus = -1.0001 mm, dXYds = 0.0000 mm dXY = -1.0001 mm VT = -76.580 uVs 17:22:13 Integrated Voltage Measurement... 17:22:22 Voltage Offset Correction: v_off_ini = 1807.838 uV, v_off_fin = 1822.661 uV 17:22:22 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 112.031 uVs 17:22:22 Integrated Voltage Measurement... 17:22:31 Voltage Offset Correction: v_off_ini = 1789.947 uV, v_off_fin = 1772.723 uV 17:22:31 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -104.989 uVs 17:22:31 Integrated Voltage Measurement... 17:22:40 Voltage Offset Correction: v_off_ini = 1784.918 uV, v_off_fin = 1796.367 uV 17:22:40 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 115.728 uVs 17:22:40 Integrated Voltage Measurement... 17:22:49 Voltage Offset Correction: v_off_ini = 1779.733 uV, v_off_fin = 1780.156 uV 17:22:49 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -106.906 uVs 17:22:49 Integrated Voltage Measurement... 17:22:58 Voltage Offset Correction: v_off_ini = 1806.760 uV, v_off_fin = 1809.596 uV 17:22:58 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 101.814 uVs 17:23:00 I2X Average and RMS of Measurements VT = -103.008 +- 6.623 uVs I2X = 9.477 +- 0.609 uTm^2 17:23:05 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 17:23:08 Integrated Voltage Measurement... 17:23:17 Voltage Offset Correction: v_off_ini = 1791.961 uV, v_off_fin = 1792.432 uV 17:23:17 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -220.762 uVs 17:23:17 Integrated Voltage Measurement... 17:23:26 Voltage Offset Correction: v_off_ini = 1777.576 uV, v_off_fin = 1805.992 uV 17:23:26 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 241.566 uVs 17:23:26 Integrated Voltage Measurement... 17:23:35 Voltage Offset Correction: v_off_ini = 1809.261 uV, v_off_fin = 1793.282 uV 17:23:35 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -223.833 uVs 17:23:35 Integrated Voltage Measurement... 17:23:44 Voltage Offset Correction: v_off_ini = 1782.927 uV, v_off_fin = 1767.191 uV 17:23:44 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 236.236 uVs 17:23:44 Integrated Voltage Measurement... 17:23:53 Voltage Offset Correction: v_off_ini = 1750.842 uV, v_off_fin = 1761.500 uV 17:23:53 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -219.866 uVs 17:23:53 Integrated Voltage Measurement... 17:24:02 Voltage Offset Correction: v_off_ini = 1786.327 uV, v_off_fin = 1767.945 uV 17:24:02 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 220.723 uVs 17:24:04 I2Y Average and RMS of Measurements VT = -227.165 +- 4.879 uVs I2Y = -20.899 +- 0.449 uTm^2 17:24:09 Wire Position: X0 = 1.0000 mm, Y0 = 0.0000 mm 17:24:13 Integrated Voltage Measurement... 17:24:22 Voltage Offset Correction: v_off_ini = 1772.844 uV, v_off_fin = 1773.604 uV 17:24:22 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -103.431 uVs 17:24:22 Integrated Voltage Measurement... 17:24:31 Voltage Offset Correction: v_off_ini = 1806.222 uV, v_off_fin = 1784.555 uV 17:24:31 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 115.756 uVs 17:24:31 Integrated Voltage Measurement... 17:24:40 Voltage Offset Correction: v_off_ini = 1778.599 uV, v_off_fin = 1792.726 uV 17:24:40 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.4999 mm dXY = -0.4999 mm VT = -108.632 uVs 17:24:40 Integrated Voltage Measurement... 17:24:49 Voltage Offset Correction: v_off_ini = 1797.738 uV, v_off_fin = 1815.131 uV 17:24:49 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = 103.617 uVs 17:24:49 Integrated Voltage Measurement... 17:24:58 Voltage Offset Correction: v_off_ini = 1802.038 uV, v_off_fin = 1780.082 uV 17:24:58 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -109.724 uVs 17:24:58 Integrated Voltage Measurement... 17:25:07 Voltage Offset Correction: v_off_ini = 1763.779 uV, v_off_fin = 1782.496 uV 17:25:07 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 120.665 uVs 17:25:11 I1X Average and RMS of Measurements VT = -110.304 +- 3.737 uVs I1X = 22.062 +- 0.746 uTm 17:25:14 Wire Position: X0 = 1.0000 mm, Y0 = 0.0000 mm 17:25:18 Integrated Voltage Measurement... 17:25:27 Voltage Offset Correction: v_off_ini = 1795.722 uV, v_off_fin = 1783.042 uV 17:25:27 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -217.170 uVs 17:25:27 Integrated Voltage Measurement... 17:25:36 Voltage Offset Correction: v_off_ini = 1797.037 uV, v_off_fin = 1827.722 uV 17:25:36 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 216.594 uVs 17:25:36 Integrated Voltage Measurement... 17:25:45 Voltage Offset Correction: v_off_ini = 1821.256 uV, v_off_fin = 1803.765 uV 17:25:45 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -217.026 uVs 17:25:45 Integrated Voltage Measurement... 17:25:54 Voltage Offset Correction: v_off_ini = 1805.507 uV, v_off_fin = 1814.724 uV 17:25:54 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 222.204 uVs 17:25:54 Integrated Voltage Measurement... 17:26:03 Voltage Offset Correction: v_off_ini = 1811.090 uV, v_off_fin = 1796.866 uV 17:26:03 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -220.864 uVs 17:26:03 Integrated Voltage Measurement... 17:26:12 Voltage Offset Correction: v_off_ini = 1807.179 uV, v_off_fin = 1843.247 uV 17:26:12 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 222.554 uVs 17:26:16 I1Y Average and RMS of Measurements VT = -219.402 +- 1.976 uVs I1Y = -43.880 +- 0.395 uTm 17:26:21 Wire Position: X0 = 1.0000 mm, Y0 = 0.0000 mm 17:26:24 Integrated Voltage Measurement... 17:26:33 Voltage Offset Correction: v_off_ini = 1815.929 uV, v_off_fin = 1820.006 uV 17:26:33 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -71.788 uVs 17:26:33 Integrated Voltage Measurement... 17:26:42 Voltage Offset Correction: v_off_ini = 1837.232 uV, v_off_fin = 1833.913 uV 17:26:42 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 82.309 uVs 17:26:42 Integrated Voltage Measurement... 17:26:51 Voltage Offset Correction: v_off_ini = 1829.272 uV, v_off_fin = 1819.205 uV 17:26:51 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -110.856 uVs 17:26:51 Integrated Voltage Measurement... 17:27:00 Voltage Offset Correction: v_off_ini = 1863.504 uV, v_off_fin = 1832.295 uV 17:27:00 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 85.026 uVs 17:27:00 Integrated Voltage Measurement... 17:27:09 Voltage Offset Correction: v_off_ini = 1823.044 uV, v_off_fin = 1822.618 uV 17:27:09 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -96.624 uVs 17:27:09 Integrated Voltage Measurement... 17:27:18 Voltage Offset Correction: v_off_ini = 1855.389 uV, v_off_fin = 1860.290 uV 17:27:18 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 82.255 uVs 17:27:20 I2X Average and RMS of Measurements VT = -88.143 +- 8.578 uVs I2X = 8.109 +- 0.789 uTm^2 17:27:25 Wire Position: X0 = 1.0000 mm, Y0 = 0.0000 mm 17:27:28 Integrated Voltage Measurement... 17:27:37 Voltage Offset Correction: v_off_ini = 1815.691 uV, v_off_fin = 1831.743 uV 17:27:37 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -235.532 uVs 17:27:37 Integrated Voltage Measurement... 17:27:46 Voltage Offset Correction: v_off_ini = 1823.339 uV, v_off_fin = 1816.825 uV 17:27:46 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 240.265 uVs 17:27:46 Integrated Voltage Measurement... 17:27:55 Voltage Offset Correction: v_off_ini = 1821.290 uV, v_off_fin = 1826.851 uV 17:27:55 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -198.090 uVs 17:27:55 Integrated Voltage Measurement... 17:28:04 Voltage Offset Correction: v_off_ini = 1819.100 uV, v_off_fin = 1826.900 uV 17:28:04 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 227.675 uVs 17:28:04 Integrated Voltage Measurement... 17:28:13 Voltage Offset Correction: v_off_ini = 1834.473 uV, v_off_fin = 1819.110 uV 17:28:13 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -236.908 uVs 17:28:13 Integrated Voltage Measurement... 17:28:22 Voltage Offset Correction: v_off_ini = 1824.028 uV, v_off_fin = 1826.138 uV 17:28:22 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 214.008 uVs 17:28:24 I2Y Average and RMS of Measurements VT = -225.413 +- 10.213 uVs I2Y = -20.738 +- 0.940 uTm^2 17:28:29 Wire Position: X0 = 2.0000 mm, Y0 = 0.0000 mm 17:28:33 Integrated Voltage Measurement... 17:28:42 Voltage Offset Correction: v_off_ini = 1831.982 uV, v_off_fin = 1832.184 uV 17:28:42 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -77.167 uVs 17:28:42 Integrated Voltage Measurement... 17:28:51 Voltage Offset Correction: v_off_ini = 1877.539 uV, v_off_fin = 1859.221 uV 17:28:51 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 50.642 uVs 17:28:51 Integrated Voltage Measurement... 17:29:00 Voltage Offset Correction: v_off_ini = 1840.289 uV, v_off_fin = 1850.596 uV 17:29:00 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -85.605 uVs 17:29:00 Integrated Voltage Measurement... 17:29:09 Voltage Offset Correction: v_off_ini = 1856.758 uV, v_off_fin = 1843.095 uV 17:29:09 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 96.213 uVs 17:29:09 Integrated Voltage Measurement... 17:29:18 Voltage Offset Correction: v_off_ini = 1867.457 uV, v_off_fin = 1853.064 uV 17:29:18 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -80.562 uVs 17:29:18 Integrated Voltage Measurement... 17:29:27 Voltage Offset Correction: v_off_ini = 1850.269 uV, v_off_fin = 1854.038 uV 17:29:27 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 69.520 uVs 17:29:31 I1X Average and RMS of Measurements VT = -76.618 +- 11.081 uVs I1X = 15.324 +- 2.216 uTm 17:29:34 Wire Position: X0 = 2.0000 mm, Y0 = 0.0000 mm 17:29:38 Integrated Voltage Measurement... 17:29:47 Voltage Offset Correction: v_off_ini = 1866.936 uV, v_off_fin = 1864.770 uV 17:29:47 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -255.978 uVs 17:29:47 Integrated Voltage Measurement... 17:29:56 Voltage Offset Correction: v_off_ini = 1898.068 uV, v_off_fin = 1865.969 uV 17:29:56 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 240.219 uVs 17:29:56 Integrated Voltage Measurement... 17:30:05 Voltage Offset Correction: v_off_ini = 1884.294 uV, v_off_fin = 1882.608 uV 17:30:05 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -222.027 uVs 17:30:05 Integrated Voltage Measurement... 17:30:14 Voltage Offset Correction: v_off_ini = 1881.204 uV, v_off_fin = 1879.266 uV 17:30:14 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 227.051 uVs 17:30:14 Integrated Voltage Measurement... 17:30:23 Voltage Offset Correction: v_off_ini = 1872.893 uV, v_off_fin = 1862.404 uV 17:30:23 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -220.622 uVs 17:30:23 Integrated Voltage Measurement... 17:30:32 Voltage Offset Correction: v_off_ini = 1893.125 uV, v_off_fin = 1884.274 uV 17:30:32 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 222.543 uVs 17:30:36 I1Y Average and RMS of Measurements VT = -231.407 +- 11.865 uVs I1Y = -46.281 +- 2.373 uTm 17:30:41 Wire Position: X0 = 2.0000 mm, Y0 = 0.0000 mm 17:30:44 Integrated Voltage Measurement... 17:30:53 Voltage Offset Correction: v_off_ini = 1862.155 uV, v_off_fin = 1848.584 uV 17:30:53 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -76.566 uVs 17:30:53 Integrated Voltage Measurement... 17:31:02 Voltage Offset Correction: v_off_ini = 1865.590 uV, v_off_fin = 1863.121 uV 17:31:02 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 53.428 uVs 17:31:02 Integrated Voltage Measurement... 17:31:11 Voltage Offset Correction: v_off_ini = 1888.101 uV, v_off_fin = 1869.767 uV 17:31:11 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -54.653 uVs 17:31:11 Integrated Voltage Measurement... 17:31:20 Voltage Offset Correction: v_off_ini = 1858.211 uV, v_off_fin = 1887.785 uV 17:31:20 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 63.872 uVs 17:31:20 Integrated Voltage Measurement... 17:31:29 Voltage Offset Correction: v_off_ini = 1878.370 uV, v_off_fin = 1864.705 uV 17:31:29 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -35.510 uVs 17:31:29 Integrated Voltage Measurement... 17:31:38 Voltage Offset Correction: v_off_ini = 1852.231 uV, v_off_fin = 1874.120 uV 17:31:38 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 87.360 uVs 17:31:40 I2X Average and RMS of Measurements VT = -61.898 +- 2.364 uVs I2X = 5.695 +- 0.217 uTm^2 17:31:45 Wire Position: X0 = 2.0000 mm, Y0 = 0.0000 mm 17:31:48 Integrated Voltage Measurement... 17:31:57 Voltage Offset Correction: v_off_ini = 1886.066 uV, v_off_fin = 1875.995 uV 17:31:57 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -213.370 uVs 17:31:57 Integrated Voltage Measurement... 17:32:06 Voltage Offset Correction: v_off_ini = 1875.377 uV, v_off_fin = 1878.833 uV 17:32:06 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 231.543 uVs 17:32:06 Integrated Voltage Measurement... 17:32:15 Voltage Offset Correction: v_off_ini = 1890.320 uV, v_off_fin = 1875.518 uV 17:32:15 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -247.563 uVs 17:32:15 Integrated Voltage Measurement... 17:32:24 Voltage Offset Correction: v_off_ini = 1881.834 uV, v_off_fin = 1887.979 uV 17:32:24 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 250.297 uVs 17:32:24 Integrated Voltage Measurement... 17:32:33 Voltage Offset Correction: v_off_ini = 1887.898 uV, v_off_fin = 1886.438 uV 17:32:33 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -227.464 uVs 17:32:33 Integrated Voltage Measurement... 17:32:42 Voltage Offset Correction: v_off_ini = 1898.001 uV, v_off_fin = 1902.663 uV 17:32:42 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 225.909 uVs 17:32:44 I2Y Average and RMS of Measurements VT = -232.691 +- 11.612 uVs I2Y = -21.408 +- 1.068 uTm^2 17:32:52 Wire Position: X0 = 0.0000 mm, Y0 = -1.0000 mm 17:32:56 Integrated Voltage Measurement... 17:33:05 Voltage Offset Correction: v_off_ini = 1902.651 uV, v_off_fin = 1895.748 uV 17:33:05 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -123.759 uVs 17:33:05 Integrated Voltage Measurement... 17:33:14 Voltage Offset Correction: v_off_ini = 1905.716 uV, v_off_fin = 1899.288 uV 17:33:14 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 128.481 uVs 17:33:14 Integrated Voltage Measurement... 17:33:23 Voltage Offset Correction: v_off_ini = 1897.413 uV, v_off_fin = 1915.544 uV 17:33:23 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -122.892 uVs 17:33:23 Integrated Voltage Measurement... 17:33:32 Voltage Offset Correction: v_off_ini = 1899.754 uV, v_off_fin = 1887.018 uV 17:33:32 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 129.538 uVs 17:33:32 Integrated Voltage Measurement... 17:33:41 Voltage Offset Correction: v_off_ini = 1908.982 uV, v_off_fin = 1895.520 uV 17:33:41 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -117.744 uVs 17:33:41 Integrated Voltage Measurement... 17:33:50 Voltage Offset Correction: v_off_ini = 1904.533 uV, v_off_fin = 1892.626 uV 17:33:50 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 151.469 uVs 17:33:54 I1X Average and RMS of Measurements VT = -128.981 +- 3.978 uVs I1X = 25.797 +- 0.795 uTm 17:33:57 Wire Position: X0 = 0.0000 mm, Y0 = -1.0000 mm 17:34:01 Integrated Voltage Measurement... 17:34:10 Voltage Offset Correction: v_off_ini = 1882.927 uV, v_off_fin = 1872.485 uV 17:34:10 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -249.001 uVs 17:34:10 Integrated Voltage Measurement... 17:34:19 Voltage Offset Correction: v_off_ini = 1905.205 uV, v_off_fin = 1908.802 uV 17:34:19 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 236.670 uVs 17:34:19 Integrated Voltage Measurement... 17:34:28 Voltage Offset Correction: v_off_ini = 1899.388 uV, v_off_fin = 1872.307 uV 17:34:28 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -245.424 uVs 17:34:28 Integrated Voltage Measurement... 17:34:37 Voltage Offset Correction: v_off_ini = 1875.116 uV, v_off_fin = 1906.763 uV 17:34:37 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 247.448 uVs 17:34:37 Integrated Voltage Measurement... 17:34:46 Voltage Offset Correction: v_off_ini = 1900.215 uV, v_off_fin = 1877.502 uV 17:34:46 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -256.343 uVs 17:34:46 Integrated Voltage Measurement... 17:34:55 Voltage Offset Correction: v_off_ini = 1902.388 uV, v_off_fin = 1897.830 uV 17:34:55 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 255.048 uVs 17:34:59 I1Y Average and RMS of Measurements VT = -248.322 +- 5.417 uVs I1Y = -49.664 +- 1.083 uTm 17:35:02 Wire Position: X0 = 0.0000 mm, Y0 = -1.0000 mm 17:35:04 Integrated Voltage Measurement... 17:35:13 Voltage Offset Correction: v_off_ini = 1902.910 uV, v_off_fin = 1900.657 uV 17:35:13 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -110.263 uVs 17:35:13 Integrated Voltage Measurement... 17:35:22 Voltage Offset Correction: v_off_ini = 1886.137 uV, v_off_fin = 1911.878 uV 17:35:22 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 96.272 uVs 17:35:22 Integrated Voltage Measurement... 17:35:31 Voltage Offset Correction: v_off_ini = 1921.393 uV, v_off_fin = 1896.971 uV 17:35:31 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -115.469 uVs 17:35:31 Integrated Voltage Measurement... 17:35:40 Voltage Offset Correction: v_off_ini = 1902.556 uV, v_off_fin = 1891.391 uV 17:35:40 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 118.883 uVs 17:35:40 Integrated Voltage Measurement... 17:35:49 Voltage Offset Correction: v_off_ini = 1881.157 uV, v_off_fin = 1876.284 uV 17:35:49 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -97.243 uVs 17:35:49 Integrated Voltage Measurement... 17:35:58 Voltage Offset Correction: v_off_ini = 1885.382 uV, v_off_fin = 1887.897 uV 17:35:58 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 114.522 uVs 17:36:01 I2X Average and RMS of Measurements VT = -108.775 +- 6.035 uVs I2X = 10.007 +- 0.555 uTm^2 17:36:03 Wire Position: X0 = 0.0000 mm, Y0 = -1.0000 mm 17:36:06 Integrated Voltage Measurement... 17:36:15 Voltage Offset Correction: v_off_ini = 1894.591 uV, v_off_fin = 1882.467 uV 17:36:15 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -259.611 uVs 17:36:15 Integrated Voltage Measurement... 17:36:24 Voltage Offset Correction: v_off_ini = 1880.957 uV, v_off_fin = 1854.900 uV 17:36:24 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 252.415 uVs 17:36:24 Integrated Voltage Measurement... 17:36:33 Voltage Offset Correction: v_off_ini = 1875.040 uV, v_off_fin = 1891.806 uV 17:36:33 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -230.085 uVs 17:36:33 Integrated Voltage Measurement... 17:36:42 Voltage Offset Correction: v_off_ini = 1907.960 uV, v_off_fin = 1891.245 uV 17:36:42 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 229.410 uVs 17:36:42 Integrated Voltage Measurement... 17:36:51 Voltage Offset Correction: v_off_ini = 1900.321 uV, v_off_fin = 1875.504 uV 17:36:51 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -244.786 uVs 17:36:51 Integrated Voltage Measurement... 17:37:00 Voltage Offset Correction: v_off_ini = 1864.938 uV, v_off_fin = 1859.346 uV 17:37:00 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 249.790 uVs 17:37:02 I2Y Average and RMS of Measurements VT = -244.349 +- 10.922 uVs I2Y = -22.480 +- 1.005 uTm^2 17:37:06 Wire Position: X0 = 0.0000 mm, Y0 = -0.5000 mm 17:37:10 Integrated Voltage Measurement... 17:37:19 Voltage Offset Correction: v_off_ini = 1868.926 uV, v_off_fin = 1873.606 uV 17:37:19 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -126.388 uVs 17:37:19 Integrated Voltage Measurement... 17:37:28 Voltage Offset Correction: v_off_ini = 1877.068 uV, v_off_fin = 1878.437 uV 17:37:28 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 109.886 uVs 17:37:28 Integrated Voltage Measurement... 17:37:37 Voltage Offset Correction: v_off_ini = 1891.633 uV, v_off_fin = 1880.753 uV 17:37:37 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -129.963 uVs 17:37:37 Integrated Voltage Measurement... 17:37:46 Voltage Offset Correction: v_off_ini = 1859.722 uV, v_off_fin = 1880.314 uV 17:37:46 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 109.551 uVs 17:37:46 Integrated Voltage Measurement... 17:37:55 Voltage Offset Correction: v_off_ini = 1896.281 uV, v_off_fin = 1876.894 uV 17:37:55 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.4999 mm dXY = -0.4999 mm VT = -138.601 uVs 17:37:55 Integrated Voltage Measurement... 17:38:04 Voltage Offset Correction: v_off_ini = 1894.216 uV, v_off_fin = 1885.088 uV 17:38:04 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 105.887 uVs 17:38:09 I1X Average and RMS of Measurements VT = -120.046 +- 1.689 uVs I1X = 24.010 +- 0.339 uTm 17:38:11 Wire Position: X0 = 0.0000 mm, Y0 = -0.5000 mm 17:38:15 Integrated Voltage Measurement... 17:38:24 Voltage Offset Correction: v_off_ini = 1891.501 uV, v_off_fin = 1900.035 uV 17:38:24 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -203.820 uVs 17:38:24 Integrated Voltage Measurement... 17:38:33 Voltage Offset Correction: v_off_ini = 1914.487 uV, v_off_fin = 1928.190 uV 17:38:33 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 220.262 uVs 17:38:33 Integrated Voltage Measurement... 17:38:42 Voltage Offset Correction: v_off_ini = 1910.210 uV, v_off_fin = 1885.345 uV 17:38:42 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -239.704 uVs 17:38:42 Integrated Voltage Measurement... 17:38:51 Voltage Offset Correction: v_off_ini = 1910.198 uV, v_off_fin = 1916.458 uV 17:38:51 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 230.318 uVs 17:38:51 Integrated Voltage Measurement... 17:39:00 Voltage Offset Correction: v_off_ini = 1912.045 uV, v_off_fin = 1894.990 uV 17:39:00 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -241.596 uVs 17:39:00 Integrated Voltage Measurement... 17:39:09 Voltage Offset Correction: v_off_ini = 1910.184 uV, v_off_fin = 1931.133 uV 17:39:09 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 238.758 uVs 17:39:13 I1Y Average and RMS of Measurements VT = -229.076 +- 12.229 uVs I1Y = -45.815 +- 2.446 uTm 17:39:16 Wire Position: X0 = 0.0000 mm, Y0 = -0.5000 mm 17:39:18 Integrated Voltage Measurement... 17:39:27 Voltage Offset Correction: v_off_ini = 1912.472 uV, v_off_fin = 1886.328 uV 17:39:27 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -118.700 uVs 17:39:27 Integrated Voltage Measurement... 17:39:36 Voltage Offset Correction: v_off_ini = 1912.493 uV, v_off_fin = 1922.157 uV 17:39:36 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 106.228 uVs 17:39:36 Integrated Voltage Measurement... 17:39:45 Voltage Offset Correction: v_off_ini = 1922.334 uV, v_off_fin = 1902.569 uV 17:39:45 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -92.993 uVs 17:39:45 Integrated Voltage Measurement... 17:39:54 Voltage Offset Correction: v_off_ini = 1902.909 uV, v_off_fin = 1923.109 uV 17:39:54 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 96.673 uVs 17:39:54 Integrated Voltage Measurement... 17:40:03 Voltage Offset Correction: v_off_ini = 1932.901 uV, v_off_fin = 1926.950 uV 17:40:03 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -110.887 uVs 17:40:03 Integrated Voltage Measurement... 17:40:12 Voltage Offset Correction: v_off_ini = 1921.784 uV, v_off_fin = 1920.073 uV 17:40:12 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 117.576 uVs 17:40:15 I2X Average and RMS of Measurements VT = -107.176 +- 8.758 uVs I2X = 9.860 +- 0.806 uTm^2 17:40:17 Wire Position: X0 = 0.0000 mm, Y0 = -0.5000 mm 17:40:20 Integrated Voltage Measurement... 17:40:29 Voltage Offset Correction: v_off_ini = 1917.366 uV, v_off_fin = 1928.120 uV 17:40:29 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -234.864 uVs 17:40:29 Integrated Voltage Measurement... 17:40:38 Voltage Offset Correction: v_off_ini = 1920.421 uV, v_off_fin = 1919.477 uV 17:40:38 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 242.051 uVs 17:40:38 Integrated Voltage Measurement... 17:40:47 Voltage Offset Correction: v_off_ini = 1930.588 uV, v_off_fin = 1937.390 uV 17:40:47 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -243.902 uVs 17:40:47 Integrated Voltage Measurement... 17:40:56 Voltage Offset Correction: v_off_ini = 1940.489 uV, v_off_fin = 1933.646 uV 17:40:56 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 236.326 uVs 17:40:56 Integrated Voltage Measurement... 17:41:05 Voltage Offset Correction: v_off_ini = 1928.199 uV, v_off_fin = 1919.982 uV 17:41:05 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -226.745 uVs 17:41:05 Integrated Voltage Measurement... 17:41:14 Voltage Offset Correction: v_off_ini = 1922.226 uV, v_off_fin = 1954.184 uV 17:41:14 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 235.816 uVs 17:41:16 I2Y Average and RMS of Measurements VT = -236.617 +- 3.834 uVs I2Y = -21.769 +- 0.353 uTm^2 17:41:20 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 17:41:25 Integrated Voltage Measurement... 17:41:34 Voltage Offset Correction: v_off_ini = 1921.169 uV, v_off_fin = 1941.374 uV 17:41:34 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4998 mm dXY = -0.4999 mm VT = -100.166 uVs 17:41:34 Integrated Voltage Measurement... 17:41:43 Voltage Offset Correction: v_off_ini = 1951.321 uV, v_off_fin = 1937.314 uV 17:41:43 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 124.778 uVs 17:41:43 Integrated Voltage Measurement... 17:41:52 Voltage Offset Correction: v_off_ini = 1935.629 uV, v_off_fin = 1936.223 uV 17:41:52 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -132.598 uVs 17:41:52 Integrated Voltage Measurement... 17:42:01 Voltage Offset Correction: v_off_ini = 1933.884 uV, v_off_fin = 1934.830 uV 17:42:01 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 141.668 uVs 17:42:01 Integrated Voltage Measurement... 17:42:10 Voltage Offset Correction: v_off_ini = 1933.970 uV, v_off_fin = 1934.107 uV 17:42:10 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -125.369 uVs 17:42:10 Integrated Voltage Measurement... 17:42:19 Voltage Offset Correction: v_off_ini = 1939.341 uV, v_off_fin = 1937.649 uV 17:42:19 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 117.565 uVs 17:42:23 I1X Average and RMS of Measurements VT = -123.691 +- 10.190 uVs I1X = 24.739 +- 2.038 uTm 17:42:25 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 17:42:29 Integrated Voltage Measurement... 17:42:38 Voltage Offset Correction: v_off_ini = 1968.684 uV, v_off_fin = 1944.287 uV 17:42:38 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.4999 mm dXY = 0.4999 mm VT = -226.496 uVs 17:42:38 Integrated Voltage Measurement... 17:42:47 Voltage Offset Correction: v_off_ini = 1940.088 uV, v_off_fin = 1943.917 uV 17:42:47 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 206.134 uVs 17:42:47 Integrated Voltage Measurement... 17:42:56 Voltage Offset Correction: v_off_ini = 1927.003 uV, v_off_fin = 1922.804 uV 17:42:56 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -225.061 uVs 17:42:56 Integrated Voltage Measurement... 17:43:05 Voltage Offset Correction: v_off_ini = 1954.899 uV, v_off_fin = 1950.141 uV 17:43:05 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 205.976 uVs 17:43:05 Integrated Voltage Measurement... 17:43:14 Voltage Offset Correction: v_off_ini = 1930.673 uV, v_off_fin = 1924.272 uV 17:43:14 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -208.415 uVs 17:43:14 Integrated Voltage Measurement... 17:43:23 Voltage Offset Correction: v_off_ini = 1918.753 uV, v_off_fin = 1937.303 uV 17:43:23 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 223.892 uVs 17:43:27 I1Y Average and RMS of Measurements VT = -215.996 +- 0.344 uVs I1Y = -43.201 +- 0.070 uTm 17:43:30 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 17:43:33 Integrated Voltage Measurement... 17:43:42 Voltage Offset Correction: v_off_ini = 1931.457 uV, v_off_fin = 1940.588 uV 17:43:42 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -93.694 uVs 17:43:42 Integrated Voltage Measurement... 17:43:51 Voltage Offset Correction: v_off_ini = 1936.662 uV, v_off_fin = 1945.939 uV 17:43:51 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 116.629 uVs 17:43:51 Integrated Voltage Measurement... 17:44:00 Voltage Offset Correction: v_off_ini = 1936.526 uV, v_off_fin = 1923.524 uV 17:44:00 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -109.913 uVs 17:44:00 Integrated Voltage Measurement... 17:44:09 Voltage Offset Correction: v_off_ini = 1964.271 uV, v_off_fin = 1953.086 uV 17:44:09 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 107.966 uVs 17:44:09 Integrated Voltage Measurement... 17:44:18 Voltage Offset Correction: v_off_ini = 1961.578 uV, v_off_fin = 1943.964 uV 17:44:18 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -108.268 uVs 17:44:18 Integrated Voltage Measurement... 17:44:27 Voltage Offset Correction: v_off_ini = 1948.771 uV, v_off_fin = 1961.269 uV 17:44:27 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 114.422 uVs 17:44:29 I2X Average and RMS of Measurements VT = -108.482 +- 2.545 uVs I2X = 9.980 +- 0.234 uTm^2 17:44:31 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 17:44:34 Integrated Voltage Measurement... 17:44:43 Voltage Offset Correction: v_off_ini = 1959.509 uV, v_off_fin = 1956.365 uV 17:44:43 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -207.406 uVs 17:44:43 Integrated Voltage Measurement... 17:44:52 Voltage Offset Correction: v_off_ini = 1958.982 uV, v_off_fin = 1958.037 uV 17:44:52 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 217.089 uVs 17:44:52 Integrated Voltage Measurement... 17:45:01 Voltage Offset Correction: v_off_ini = 1958.599 uV, v_off_fin = 1950.701 uV 17:45:01 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -206.127 uVs 17:45:01 Integrated Voltage Measurement... 17:45:10 Voltage Offset Correction: v_off_ini = 1931.730 uV, v_off_fin = 1934.326 uV 17:45:10 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 236.609 uVs 17:45:10 Integrated Voltage Measurement... 17:45:19 Voltage Offset Correction: v_off_ini = 1918.063 uV, v_off_fin = 1924.326 uV 17:45:19 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -205.620 uVs 17:45:19 Integrated Voltage Measurement... 17:45:28 Voltage Offset Correction: v_off_ini = 1915.328 uV, v_off_fin = 1950.295 uV 17:45:28 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 217.183 uVs 17:45:30 I2Y Average and RMS of Measurements VT = -215.006 +- 4.512 uVs I2Y = -19.781 +- 0.415 uTm^2 17:45:34 Wire Position: X0 = 0.0000 mm, Y0 = 0.5000 mm 17:45:39 Integrated Voltage Measurement... 17:45:48 Voltage Offset Correction: v_off_ini = 1945.137 uV, v_off_fin = 1953.030 uV 17:45:48 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -103.924 uVs 17:45:48 Integrated Voltage Measurement... 17:45:57 Voltage Offset Correction: v_off_ini = 1952.478 uV, v_off_fin = 1952.132 uV 17:45:57 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 126.247 uVs 17:45:57 Integrated Voltage Measurement... 17:46:06 Voltage Offset Correction: v_off_ini = 1919.965 uV, v_off_fin = 1950.418 uV 17:46:06 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -121.592 uVs 17:46:06 Integrated Voltage Measurement... 17:46:15 Voltage Offset Correction: v_off_ini = 1941.096 uV, v_off_fin = 1949.698 uV 17:46:15 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 112.123 uVs 17:46:15 Integrated Voltage Measurement... 17:46:24 Voltage Offset Correction: v_off_ini = 1957.757 uV, v_off_fin = 1953.993 uV 17:46:24 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -112.552 uVs 17:46:24 Integrated Voltage Measurement... 17:46:33 Voltage Offset Correction: v_off_ini = 1960.386 uV, v_off_fin = 1958.506 uV 17:46:33 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 114.829 uVs 17:46:37 I1X Average and RMS of Measurements VT = -115.211 +- 1.296 uVs I1X = 23.043 +- 0.259 uTm 17:46:39 Wire Position: X0 = 0.0000 mm, Y0 = 0.5000 mm 17:46:44 Integrated Voltage Measurement... 17:46:52 Voltage Offset Correction: v_off_ini = 1936.718 uV, v_off_fin = 1938.478 uV 17:46:53 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -210.794 uVs 17:46:53 Integrated Voltage Measurement... 17:47:01 Voltage Offset Correction: v_off_ini = 1935.397 uV, v_off_fin = 1928.993 uV 17:47:01 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 204.829 uVs 17:47:02 Integrated Voltage Measurement... 17:47:10 Voltage Offset Correction: v_off_ini = 1944.227 uV, v_off_fin = 1930.573 uV 17:47:10 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -206.760 uVs 17:47:10 Integrated Voltage Measurement... 17:47:19 Voltage Offset Correction: v_off_ini = 1928.084 uV, v_off_fin = 1952.101 uV 17:47:19 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 216.316 uVs 17:47:19 Integrated Voltage Measurement... 17:47:28 Voltage Offset Correction: v_off_ini = 1926.630 uV, v_off_fin = 1928.683 uV 17:47:28 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -188.615 uVs 17:47:28 Integrated Voltage Measurement... 17:47:37 Voltage Offset Correction: v_off_ini = 1936.059 uV, v_off_fin = 1953.717 uV 17:47:37 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 197.269 uVs 17:47:42 I1Y Average and RMS of Measurements VT = -204.097 +- 8.033 uVs I1Y = -40.817 +- 1.605 uTm 17:47:44 Wire Position: X0 = 0.0000 mm, Y0 = 0.5000 mm 17:47:47 Integrated Voltage Measurement... 17:47:56 Voltage Offset Correction: v_off_ini = 1923.903 uV, v_off_fin = 1912.271 uV 17:47:56 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -111.008 uVs 17:47:56 Integrated Voltage Measurement... 17:48:05 Voltage Offset Correction: v_off_ini = 1937.084 uV, v_off_fin = 1931.929 uV 17:48:05 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 83.881 uVs 17:48:05 Integrated Voltage Measurement... 17:48:14 Voltage Offset Correction: v_off_ini = 1941.125 uV, v_off_fin = 1944.665 uV 17:48:14 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -102.769 uVs 17:48:14 Integrated Voltage Measurement... 17:48:23 Voltage Offset Correction: v_off_ini = 1936.522 uV, v_off_fin = 1931.469 uV 17:48:23 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 91.567 uVs 17:48:23 Integrated Voltage Measurement... 17:48:32 Voltage Offset Correction: v_off_ini = 1931.721 uV, v_off_fin = 1933.306 uV 17:48:32 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -92.279 uVs 17:48:32 Integrated Voltage Measurement... 17:48:41 Voltage Offset Correction: v_off_ini = 1943.781 uV, v_off_fin = 1963.143 uV 17:48:41 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 86.238 uVs 17:48:43 I2X Average and RMS of Measurements VT = -94.624 +- 3.795 uVs I2X = 8.705 +- 0.349 uTm^2 17:48:46 Wire Position: X0 = 0.0000 mm, Y0 = 0.5000 mm 17:48:48 Integrated Voltage Measurement... 17:48:57 Voltage Offset Correction: v_off_ini = 1943.015 uV, v_off_fin = 1943.588 uV 17:48:57 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -197.533 uVs 17:48:57 Integrated Voltage Measurement... 17:49:06 Voltage Offset Correction: v_off_ini = 1985.491 uV, v_off_fin = 1987.129 uV 17:49:06 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 184.647 uVs 17:49:06 Integrated Voltage Measurement... 17:49:15 Voltage Offset Correction: v_off_ini = 1984.608 uV, v_off_fin = 1973.243 uV 17:49:15 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -220.972 uVs 17:49:15 Integrated Voltage Measurement... 17:49:24 Voltage Offset Correction: v_off_ini = 1970.006 uV, v_off_fin = 1950.338 uV 17:49:24 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 171.358 uVs 17:49:24 Integrated Voltage Measurement... 17:49:33 Voltage Offset Correction: v_off_ini = 1952.244 uV, v_off_fin = 1952.133 uV 17:49:33 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -212.194 uVs 17:49:33 Integrated Voltage Measurement... 17:49:42 Voltage Offset Correction: v_off_ini = 1962.420 uV, v_off_fin = 1941.007 uV 17:49:42 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 143.194 uVs 17:49:45 I2Y Average and RMS of Measurements VT = -188.316 +- 7.792 uVs I2Y = -17.325 +- 0.717 uTm^2 17:49:49 Wire Position: X0 = 0.0000 mm, Y0 = 1.0000 mm 17:49:53 Integrated Voltage Measurement... 17:50:02 Voltage Offset Correction: v_off_ini = 1975.031 uV, v_off_fin = 1957.344 uV 17:50:02 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -92.828 uVs 17:50:02 Integrated Voltage Measurement... 17:50:11 Voltage Offset Correction: v_off_ini = 1952.295 uV, v_off_fin = 1950.488 uV 17:50:11 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 124.415 uVs 17:50:11 Integrated Voltage Measurement... 17:50:20 Voltage Offset Correction: v_off_ini = 1959.041 uV, v_off_fin = 1976.781 uV 17:50:20 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -131.724 uVs 17:50:20 Integrated Voltage Measurement... 17:50:29 Voltage Offset Correction: v_off_ini = 1947.681 uV, v_off_fin = 1950.878 uV 17:50:29 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 104.026 uVs 17:50:29 Integrated Voltage Measurement... 17:50:38 Voltage Offset Correction: v_off_ini = 1961.396 uV, v_off_fin = 1956.764 uV 17:50:38 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -105.598 uVs 17:50:38 Integrated Voltage Measurement... 17:50:47 Voltage Offset Correction: v_off_ini = 1939.264 uV, v_off_fin = 1947.400 uV 17:50:47 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 119.535 uVs 17:50:51 I1X Average and RMS of Measurements VT = -113.021 +- 3.791 uVs I1X = 22.604 +- 0.758 uTm 17:50:53 Wire Position: X0 = 0.0000 mm, Y0 = 1.0000 mm 17:50:58 Integrated Voltage Measurement... 17:51:07 Voltage Offset Correction: v_off_ini = 1944.375 uV, v_off_fin = 1949.225 uV 17:51:07 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -181.640 uVs 17:51:07 Integrated Voltage Measurement... 17:51:16 Voltage Offset Correction: v_off_ini = 1966.798 uV, v_off_fin = 1987.960 uV 17:51:16 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 191.259 uVs 17:51:16 Integrated Voltage Measurement... 17:51:25 Voltage Offset Correction: v_off_ini = 1950.441 uV, v_off_fin = 1962.105 uV 17:51:25 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -187.652 uVs 17:51:25 Integrated Voltage Measurement... 17:51:34 Voltage Offset Correction: v_off_ini = 1944.953 uV, v_off_fin = 1961.447 uV 17:51:34 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 210.316 uVs 17:51:34 Integrated Voltage Measurement... 17:51:43 Voltage Offset Correction: v_off_ini = 1937.388 uV, v_off_fin = 1929.111 uV 17:51:43 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -204.387 uVs 17:51:43 Integrated Voltage Measurement... 17:51:52 Voltage Offset Correction: v_off_ini = 1950.752 uV, v_off_fin = 1953.198 uV 17:51:52 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 184.811 uVs 17:51:56 I1Y Average and RMS of Measurements VT = -193.344 +- 5.194 uVs I1Y = -38.666 +- 1.039 uTm 17:51:58 Wire Position: X0 = 0.0000 mm, Y0 = 1.0000 mm 17:52:01 Integrated Voltage Measurement... 17:52:10 Voltage Offset Correction: v_off_ini = 1946.639 uV, v_off_fin = 1950.540 uV 17:52:10 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -79.359 uVs 17:52:10 Integrated Voltage Measurement... 17:52:19 Voltage Offset Correction: v_off_ini = 1930.541 uV, v_off_fin = 1953.753 uV 17:52:19 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 84.203 uVs 17:52:19 Integrated Voltage Measurement... 17:52:28 Voltage Offset Correction: v_off_ini = 1968.462 uV, v_off_fin = 1944.040 uV 17:52:28 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -110.927 uVs 17:52:28 Integrated Voltage Measurement... 17:52:37 Voltage Offset Correction: v_off_ini = 1955.514 uV, v_off_fin = 1954.564 uV 17:52:37 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 109.848 uVs 17:52:37 Integrated Voltage Measurement... 17:52:46 Voltage Offset Correction: v_off_ini = 1944.400 uV, v_off_fin = 1939.716 uV 17:52:46 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -77.469 uVs 17:52:46 Integrated Voltage Measurement... 17:52:55 Voltage Offset Correction: v_off_ini = 1951.638 uV, v_off_fin = 1970.035 uV 17:52:55 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 94.409 uVs 17:52:57 I2X Average and RMS of Measurements VT = -92.702 +- 12.620 uVs I2X = 8.529 +- 1.161 uTm^2 17:53:00 Wire Position: X0 = 0.0000 mm, Y0 = 1.0000 mm 17:53:02 Integrated Voltage Measurement... 17:53:11 Voltage Offset Correction: v_off_ini = 1973.271 uV, v_off_fin = 1987.034 uV 17:53:11 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -207.090 uVs 17:53:11 Integrated Voltage Measurement... 17:53:20 Voltage Offset Correction: v_off_ini = 1983.599 uV, v_off_fin = 1995.446 uV 17:53:20 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 206.430 uVs 17:53:20 Integrated Voltage Measurement... 17:53:29 Voltage Offset Correction: v_off_ini = 1994.498 uV, v_off_fin = 1994.368 uV 17:53:29 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -174.870 uVs 17:53:29 Integrated Voltage Measurement... 17:53:38 Voltage Offset Correction: v_off_ini = 1971.300 uV, v_off_fin = 1997.093 uV 17:53:38 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 198.952 uVs 17:53:38 Integrated Voltage Measurement... 17:53:47 Voltage Offset Correction: v_off_ini = 1996.854 uV, v_off_fin = 1986.387 uV 17:53:47 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -183.762 uVs 17:53:47 Integrated Voltage Measurement... 17:53:56 Voltage Offset Correction: v_off_ini = 1980.761 uV, v_off_fin = 2006.664 uV 17:53:56 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 174.572 uVs 17:53:59 I2Y Average and RMS of Measurements VT = -190.946 +- 11.621 uVs I2Y = -17.567 +- 1.069 uTm^2 17:54:16 Wire Position: X0 = -2.0000 mm, Y0 = 0.0000 mm 17:54:20 Integrated Voltage Measurement... 17:54:29 Voltage Offset Correction: v_off_ini = 2029.353 uV, v_off_fin = 2005.476 uV 17:54:29 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -192.361 uVs 17:54:29 Integrated Voltage Measurement... 17:54:38 Voltage Offset Correction: v_off_ini = 2008.417 uV, v_off_fin = 2037.215 uV 17:54:38 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 181.613 uVs 17:54:38 Integrated Voltage Measurement... 17:54:47 Voltage Offset Correction: v_off_ini = 2022.417 uV, v_off_fin = 2026.098 uV 17:54:47 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -174.250 uVs 17:54:47 Integrated Voltage Measurement... 17:54:56 Voltage Offset Correction: v_off_ini = 2029.877 uV, v_off_fin = 2044.095 uV 17:54:56 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 173.642 uVs 17:54:56 Integrated Voltage Measurement... 17:55:05 Voltage Offset Correction: v_off_ini = 2030.785 uV, v_off_fin = 2027.091 uV 17:55:05 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -175.034 uVs 17:55:05 Integrated Voltage Measurement... 17:55:14 Voltage Offset Correction: v_off_ini = 2015.428 uV, v_off_fin = 2020.481 uV 17:55:14 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 159.248 uVs 17:55:18 I1X Average and RMS of Measurements VT = -176.025 +- 8.234 uVs I1X = 35.205 +- 1.647 uTm 17:55:21 Wire Position: X0 = -2.0000 mm, Y0 = 0.0000 mm 17:55:25 Integrated Voltage Measurement... 17:55:34 Voltage Offset Correction: v_off_ini = 2007.208 uV, v_off_fin = 2008.900 uV 17:55:34 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -179.176 uVs 17:55:34 Integrated Voltage Measurement... 17:55:43 Voltage Offset Correction: v_off_ini = 2025.300 uV, v_off_fin = 2018.364 uV 17:55:43 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 182.611 uVs 17:55:43 Integrated Voltage Measurement... 17:55:52 Voltage Offset Correction: v_off_ini = 2018.911 uV, v_off_fin = 2031.659 uV 17:55:52 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -181.124 uVs 17:55:52 Integrated Voltage Measurement... 17:56:01 Voltage Offset Correction: v_off_ini = 2038.984 uV, v_off_fin = 2056.374 uV 17:56:01 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 182.450 uVs 17:56:01 Integrated Voltage Measurement... 17:56:10 Voltage Offset Correction: v_off_ini = 2042.732 uV, v_off_fin = 2039.734 uV 17:56:10 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -168.977 uVs 17:56:10 Integrated Voltage Measurement... 17:56:19 Voltage Offset Correction: v_off_ini = 2033.444 uV, v_off_fin = 2039.499 uV 17:56:19 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 184.177 uVs 17:56:23 I1Y Average and RMS of Measurements VT = -179.753 +- 2.275 uVs I1Y = -35.951 +- 0.455 uTm 17:56:28 Wire Position: X0 = -2.0000 mm, Y0 = 0.0000 mm 17:56:30 Integrated Voltage Measurement... 17:56:39 Voltage Offset Correction: v_off_ini = 2025.519 uV, v_off_fin = 1990.195 uV 17:56:39 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -171.993 uVs 17:56:39 Integrated Voltage Measurement... 17:56:48 Voltage Offset Correction: v_off_ini = 2020.900 uV, v_off_fin = 2040.936 uV 17:56:48 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 170.383 uVs 17:56:48 Integrated Voltage Measurement... 17:56:57 Voltage Offset Correction: v_off_ini = 2022.970 uV, v_off_fin = 2014.873 uV 17:56:57 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -168.520 uVs 17:56:58 Integrated Voltage Measurement... 17:57:06 Voltage Offset Correction: v_off_ini = 2052.114 uV, v_off_fin = 2055.017 uV 17:57:06 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 145.693 uVs 17:57:07 Integrated Voltage Measurement... 17:57:15 Voltage Offset Correction: v_off_ini = 2052.833 uV, v_off_fin = 2053.291 uV 17:57:15 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -161.550 uVs 17:57:16 Integrated Voltage Measurement... 17:57:24 Voltage Offset Correction: v_off_ini = 2043.101 uV, v_off_fin = 2051.009 uV 17:57:25 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 173.840 uVs 17:57:27 I2X Average and RMS of Measurements VT = -165.330 +- 5.987 uVs I2X = 15.210 +- 0.551 uTm^2 17:57:32 Wire Position: X0 = -2.0000 mm, Y0 = 0.0000 mm 17:57:35 Integrated Voltage Measurement... 17:57:44 Voltage Offset Correction: v_off_ini = 2023.559 uV, v_off_fin = 2026.178 uV 17:57:44 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -183.227 uVs 17:57:44 Integrated Voltage Measurement... 17:57:53 Voltage Offset Correction: v_off_ini = 2040.167 uV, v_off_fin = 2033.939 uV 17:57:53 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 168.052 uVs 17:57:53 Integrated Voltage Measurement... 17:58:02 Voltage Offset Correction: v_off_ini = 2054.256 uV, v_off_fin = 2034.851 uV 17:58:02 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -185.489 uVs 17:58:02 Integrated Voltage Measurement... 17:58:11 Voltage Offset Correction: v_off_ini = 2034.622 uV, v_off_fin = 2042.868 uV 17:58:11 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 170.885 uVs 17:58:11 Integrated Voltage Measurement... 17:58:20 Voltage Offset Correction: v_off_ini = 2063.971 uV, v_off_fin = 2071.935 uV 17:58:20 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -195.674 uVs 17:58:20 Integrated Voltage Measurement... 17:58:29 Voltage Offset Correction: v_off_ini = 2052.271 uV, v_off_fin = 2044.651 uV 17:58:29 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 195.085 uVs 17:58:31 I2Y Average and RMS of Measurements VT = -183.069 +- 8.767 uVs I2Y = -16.842 +- 0.807 uTm^2 17:58:36 Wire Position: X0 = -1.0000 mm, Y0 = 0.0000 mm 17:58:40 Integrated Voltage Measurement... 17:58:49 Voltage Offset Correction: v_off_ini = 2078.385 uV, v_off_fin = 2073.465 uV 17:58:49 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -147.659 uVs 17:58:49 Integrated Voltage Measurement... 17:58:58 Voltage Offset Correction: v_off_ini = 2047.767 uV, v_off_fin = 2057.512 uV 17:58:58 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 149.996 uVs 17:58:58 Integrated Voltage Measurement... 17:59:07 Voltage Offset Correction: v_off_ini = 2048.335 uV, v_off_fin = 2053.091 uV 17:59:07 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -141.264 uVs 17:59:07 Integrated Voltage Measurement... 17:59:16 Voltage Offset Correction: v_off_ini = 2040.824 uV, v_off_fin = 2053.879 uV 17:59:16 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 150.996 uVs 17:59:16 Integrated Voltage Measurement... 17:59:25 Voltage Offset Correction: v_off_ini = 2047.152 uV, v_off_fin = 2052.681 uV 17:59:25 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -130.168 uVs 17:59:25 Integrated Voltage Measurement... 17:59:34 Voltage Offset Correction: v_off_ini = 2061.248 uV, v_off_fin = 2093.668 uV 17:59:34 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 136.693 uVs 17:59:38 I1X Average and RMS of Measurements VT = -142.796 +- 6.713 uVs I1X = 28.560 +- 1.343 uTm 17:59:40 Wire Position: X0 = -1.0000 mm, Y0 = 0.0000 mm 17:59:45 Integrated Voltage Measurement... 17:59:54 Voltage Offset Correction: v_off_ini = 2089.167 uV, v_off_fin = 2071.178 uV 17:59:54 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -166.889 uVs 17:59:54 Integrated Voltage Measurement... 18:00:03 Voltage Offset Correction: v_off_ini = 2076.504 uV, v_off_fin = 2092.030 uV 18:00:03 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 210.183 uVs 18:00:03 Integrated Voltage Measurement... 18:00:12 Voltage Offset Correction: v_off_ini = 2084.728 uV, v_off_fin = 2072.305 uV 18:00:12 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -196.375 uVs 18:00:12 Integrated Voltage Measurement... 18:00:20 Voltage Offset Correction: v_off_ini = 2086.415 uV, v_off_fin = 2074.278 uV 18:00:20 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 194.603 uVs 18:00:21 Integrated Voltage Measurement... 18:00:29 Voltage Offset Correction: v_off_ini = 2071.272 uV, v_off_fin = 2037.771 uV 18:00:29 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -182.889 uVs 18:00:29 Integrated Voltage Measurement... 18:00:38 Voltage Offset Correction: v_off_ini = 2072.722 uV, v_off_fin = 2067.955 uV 18:00:38 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 194.629 uVs 18:00:43 I1Y Average and RMS of Measurements VT = -190.928 +- 3.226 uVs I1Y = -38.186 +- 0.645 uTm 18:00:48 Wire Position: X0 = -1.0000 mm, Y0 = 0.0000 mm 18:00:50 Integrated Voltage Measurement... 18:00:59 Voltage Offset Correction: v_off_ini = 2071.454 uV, v_off_fin = 2080.300 uV 18:00:59 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -129.328 uVs 18:00:59 Integrated Voltage Measurement... 18:01:08 Voltage Offset Correction: v_off_ini = 2079.279 uV, v_off_fin = 2091.365 uV 18:01:08 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 133.801 uVs 18:01:08 Integrated Voltage Measurement... 18:01:17 Voltage Offset Correction: v_off_ini = 2084.860 uV, v_off_fin = 2094.337 uV 18:01:17 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -126.445 uVs 18:01:17 Integrated Voltage Measurement... 18:01:26 Voltage Offset Correction: v_off_ini = 2077.350 uV, v_off_fin = 2105.829 uV 18:01:26 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 140.114 uVs 18:01:26 Integrated Voltage Measurement... 18:01:35 Voltage Offset Correction: v_off_ini = 2108.308 uV, v_off_fin = 2115.417 uV 18:01:35 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -110.522 uVs 18:01:35 Integrated Voltage Measurement... 18:01:44 Voltage Offset Correction: v_off_ini = 2118.450 uV, v_off_fin = 2112.155 uV 18:01:44 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 109.940 uVs 18:01:47 I2X Average and RMS of Measurements VT = -125.025 +- 10.484 uVs I2X = 11.502 +- 0.965 uTm^2 18:01:52 Wire Position: X0 = -1.0000 mm, Y0 = 0.0000 mm 18:01:54 Integrated Voltage Measurement... 18:02:03 Voltage Offset Correction: v_off_ini = 2095.487 uV, v_off_fin = 2107.444 uV 18:02:03 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -177.863 uVs 18:02:03 Integrated Voltage Measurement... 18:02:12 Voltage Offset Correction: v_off_ini = 2108.547 uV, v_off_fin = 2113.417 uV 18:02:12 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 186.969 uVs 18:02:12 Integrated Voltage Measurement... 18:02:21 Voltage Offset Correction: v_off_ini = 2106.631 uV, v_off_fin = 2127.285 uV 18:02:21 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -183.183 uVs 18:02:21 Integrated Voltage Measurement... 18:02:30 Voltage Offset Correction: v_off_ini = 2137.116 uV, v_off_fin = 2137.004 uV 18:02:30 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 206.411 uVs 18:02:30 Integrated Voltage Measurement... 18:02:39 Voltage Offset Correction: v_off_ini = 2145.989 uV, v_off_fin = 2136.148 uV 18:02:39 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -192.511 uVs 18:02:39 Integrated Voltage Measurement... 18:02:48 Voltage Offset Correction: v_off_ini = 2127.946 uV, v_off_fin = 2134.277 uV 18:02:48 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 199.451 uVs 18:02:51 I2Y Average and RMS of Measurements VT = -191.065 +- 6.135 uVs I2Y = -17.578 +- 0.564 uTm^2 18:02:55 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 18:03:00 Integrated Voltage Measurement... 18:03:09 Voltage Offset Correction: v_off_ini = 2110.328 uV, v_off_fin = 2121.831 uV 18:03:09 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -114.124 uVs 18:03:09 Integrated Voltage Measurement... 18:03:18 Voltage Offset Correction: v_off_ini = 2129.990 uV, v_off_fin = 2133.898 uV 18:03:18 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 131.075 uVs 18:03:18 Integrated Voltage Measurement... 18:03:27 Voltage Offset Correction: v_off_ini = 2143.863 uV, v_off_fin = 2146.096 uV 18:03:27 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -116.774 uVs 18:03:27 Integrated Voltage Measurement... 18:03:36 Voltage Offset Correction: v_off_ini = 2127.319 uV, v_off_fin = 2136.854 uV 18:03:36 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 121.474 uVs 18:03:36 Integrated Voltage Measurement... 18:03:45 Voltage Offset Correction: v_off_ini = 2124.972 uV, v_off_fin = 2127.565 uV 18:03:45 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -102.967 uVs 18:03:45 Integrated Voltage Measurement... 18:03:54 Voltage Offset Correction: v_off_ini = 2144.195 uV, v_off_fin = 2138.167 uV 18:03:54 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 104.644 uVs 18:03:58 I1X Average and RMS of Measurements VT = -115.176 +- 8.164 uVs I1X = 23.035 +- 1.633 uTm 18:04:00 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 18:04:05 Integrated Voltage Measurement... 18:04:14 Voltage Offset Correction: v_off_ini = 2132.056 uV, v_off_fin = 2093.400 uV 18:04:14 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -214.629 uVs 18:04:14 Integrated Voltage Measurement... 18:04:22 Voltage Offset Correction: v_off_ini = 2111.761 uV, v_off_fin = 2124.392 uV 18:04:23 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 226.233 uVs 18:04:23 Integrated Voltage Measurement... 18:04:31 Voltage Offset Correction: v_off_ini = 2110.727 uV, v_off_fin = 2113.164 uV 18:04:32 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -206.774 uVs 18:04:32 Integrated Voltage Measurement... 18:04:40 Voltage Offset Correction: v_off_ini = 2149.624 uV, v_off_fin = 2153.150 uV 18:04:41 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 209.592 uVs 18:04:41 Integrated Voltage Measurement... 18:04:49 Voltage Offset Correction: v_off_ini = 2146.765 uV, v_off_fin = 2143.171 uV 18:04:49 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -196.486 uVs 18:04:50 Integrated Voltage Measurement... 18:04:58 Voltage Offset Correction: v_off_ini = 2159.198 uV, v_off_fin = 2166.335 uV 18:04:58 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 214.084 uVs 18:05:03 I1Y Average and RMS of Measurements VT = -211.300 +- 6.564 uVs I1Y = -42.260 +- 1.313 uTm 18:05:08 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 18:05:10 Integrated Voltage Measurement... 18:05:19 Voltage Offset Correction: v_off_ini = 2146.016 uV, v_off_fin = 2141.875 uV 18:05:19 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -84.087 uVs 18:05:19 Integrated Voltage Measurement... 18:05:28 Voltage Offset Correction: v_off_ini = 2146.977 uV, v_off_fin = 2158.605 uV 18:05:28 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 96.856 uVs 18:05:28 Integrated Voltage Measurement... 18:05:37 Voltage Offset Correction: v_off_ini = 2157.870 uV, v_off_fin = 2144.970 uV 18:05:37 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -90.453 uVs 18:05:37 Integrated Voltage Measurement... 18:05:46 Voltage Offset Correction: v_off_ini = 2151.318 uV, v_off_fin = 2156.292 uV 18:05:46 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 99.233 uVs 18:05:46 Integrated Voltage Measurement... 18:05:55 Voltage Offset Correction: v_off_ini = 2167.291 uV, v_off_fin = 2188.945 uV 18:05:55 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -103.420 uVs 18:05:55 Integrated Voltage Measurement... 18:06:04 Voltage Offset Correction: v_off_ini = 2178.975 uV, v_off_fin = 2185.615 uV 18:06:04 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 103.996 uVs 18:06:07 I2X Average and RMS of Measurements VT = -96.341 +- 5.507 uVs I2X = 8.864 +- 0.506 uTm^2 18:06:12 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 18:06:14 Integrated Voltage Measurement... 18:06:23 Voltage Offset Correction: v_off_ini = 2166.426 uV, v_off_fin = 2172.594 uV 18:06:23 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -213.170 uVs 18:06:23 Integrated Voltage Measurement... 18:06:32 Voltage Offset Correction: v_off_ini = 2177.270 uV, v_off_fin = 2170.249 uV 18:06:32 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 185.907 uVs 18:06:32 Integrated Voltage Measurement... 18:06:41 Voltage Offset Correction: v_off_ini = 2170.040 uV, v_off_fin = 2182.695 uV 18:06:41 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -167.009 uVs 18:06:41 Integrated Voltage Measurement... 18:06:50 Voltage Offset Correction: v_off_ini = 2179.039 uV, v_off_fin = 2183.702 uV 18:06:50 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 195.341 uVs 18:06:50 Integrated Voltage Measurement... 18:06:59 Voltage Offset Correction: v_off_ini = 2165.249 uV, v_off_fin = 2174.456 uV 18:06:59 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -194.349 uVs 18:06:59 Integrated Voltage Measurement... 18:07:08 Voltage Offset Correction: v_off_ini = 2164.616 uV, v_off_fin = 2171.763 uV 18:07:08 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 187.984 uVs 18:07:11 I2Y Average and RMS of Measurements VT = -190.627 +- 7.506 uVs I2Y = -17.538 +- 0.691 uTm^2 18:07:15 Wire Position: X0 = 1.0000 mm, Y0 = 0.0000 mm 18:07:20 Integrated Voltage Measurement... 18:07:29 Voltage Offset Correction: v_off_ini = 2179.694 uV, v_off_fin = 2180.790 uV 18:07:29 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -119.918 uVs 18:07:29 Integrated Voltage Measurement... 18:07:38 Voltage Offset Correction: v_off_ini = 2184.127 uV, v_off_fin = 2169.074 uV 18:07:38 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 111.872 uVs 18:07:38 Integrated Voltage Measurement... 18:07:47 Voltage Offset Correction: v_off_ini = 2154.263 uV, v_off_fin = 2154.379 uV 18:07:47 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -86.916 uVs 18:07:47 Integrated Voltage Measurement... 18:07:56 Voltage Offset Correction: v_off_ini = 2159.102 uV, v_off_fin = 2153.166 uV 18:07:56 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 94.761 uVs 18:07:56 Integrated Voltage Measurement... 18:08:05 Voltage Offset Correction: v_off_ini = 2133.524 uV, v_off_fin = 2168.922 uV 18:08:05 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -64.743 uVs 18:08:05 Integrated Voltage Measurement... 18:08:14 Voltage Offset Correction: v_off_ini = 2170.419 uV, v_off_fin = 2191.059 uV 18:08:14 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 100.448 uVs 18:08:18 I1X Average and RMS of Measurements VT = -96.443 +- 14.160 uVs I1X = 19.289 +- 2.832 uTm 18:08:20 Wire Position: X0 = 1.0000 mm, Y0 = 0.0000 mm 18:08:25 Integrated Voltage Measurement... 18:08:33 Voltage Offset Correction: v_off_ini = 2171.190 uV, v_off_fin = 2163.569 uV 18:08:33 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -215.823 uVs 18:08:34 Integrated Voltage Measurement... 18:08:42 Voltage Offset Correction: v_off_ini = 2153.780 uV, v_off_fin = 2152.230 uV 18:08:42 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 224.789 uVs 18:08:43 Integrated Voltage Measurement... 18:08:51 Voltage Offset Correction: v_off_ini = 2143.162 uV, v_off_fin = 2153.328 uV 18:08:51 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -180.570 uVs 18:08:51 Integrated Voltage Measurement... 18:09:00 Voltage Offset Correction: v_off_ini = 2143.821 uV, v_off_fin = 2153.234 uV 18:09:00 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 207.076 uVs 18:09:00 Integrated Voltage Measurement... 18:09:09 Voltage Offset Correction: v_off_ini = 2137.392 uV, v_off_fin = 2117.402 uV 18:09:09 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -221.253 uVs 18:09:09 Integrated Voltage Measurement... 18:09:18 Voltage Offset Correction: v_off_ini = 2123.996 uV, v_off_fin = 2132.154 uV 18:09:18 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 240.046 uVs 18:09:23 I1Y Average and RMS of Measurements VT = -214.926 +- 15.508 uVs I1Y = -42.985 +- 3.102 uTm 18:09:28 Wire Position: X0 = 1.0000 mm, Y0 = 0.0000 mm 18:09:30 Integrated Voltage Measurement... 18:09:39 Voltage Offset Correction: v_off_ini = 2147.568 uV, v_off_fin = 2134.581 uV 18:09:39 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -77.861 uVs 18:09:39 Integrated Voltage Measurement... 18:09:48 Voltage Offset Correction: v_off_ini = 2159.393 uV, v_off_fin = 2168.936 uV 18:09:48 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 60.001 uVs 18:09:48 Integrated Voltage Measurement... 18:09:57 Voltage Offset Correction: v_off_ini = 2166.204 uV, v_off_fin = 2143.700 uV 18:09:57 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -83.463 uVs 18:09:57 Integrated Voltage Measurement... 18:10:06 Voltage Offset Correction: v_off_ini = 2172.043 uV, v_off_fin = 2174.292 uV 18:10:06 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 58.129 uVs 18:10:06 Integrated Voltage Measurement... 18:10:15 Voltage Offset Correction: v_off_ini = 2167.613 uV, v_off_fin = 2148.426 uV 18:10:15 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -77.947 uVs 18:10:15 Integrated Voltage Measurement... 18:10:24 Voltage Offset Correction: v_off_ini = 2172.582 uV, v_off_fin = 2181.341 uV 18:10:24 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 69.893 uVs 18:10:27 I2X Average and RMS of Measurements VT = -71.216 +- 2.058 uVs I2X = 6.552 +- 0.189 uTm^2 18:10:32 Wire Position: X0 = 1.0000 mm, Y0 = 0.0000 mm 18:10:34 Integrated Voltage Measurement... 18:10:43 Voltage Offset Correction: v_off_ini = 2163.737 uV, v_off_fin = 2172.223 uV 18:10:43 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -190.161 uVs 18:10:43 Integrated Voltage Measurement... 18:10:52 Voltage Offset Correction: v_off_ini = 2168.537 uV, v_off_fin = 2171.816 uV 18:10:52 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 198.964 uVs 18:10:52 Integrated Voltage Measurement... 18:11:01 Voltage Offset Correction: v_off_ini = 2181.619 uV, v_off_fin = 2169.378 uV 18:11:01 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -216.158 uVs 18:11:01 Integrated Voltage Measurement... 18:11:10 Voltage Offset Correction: v_off_ini = 2168.529 uV, v_off_fin = 2199.275 uV 18:11:10 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 238.151 uVs 18:11:10 Integrated Voltage Measurement... 18:11:19 Voltage Offset Correction: v_off_ini = 2182.006 uV, v_off_fin = 2190.074 uV 18:11:19 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -221.735 uVs 18:11:19 Integrated Voltage Measurement... 18:11:28 Voltage Offset Correction: v_off_ini = 2157.614 uV, v_off_fin = 2166.911 uV 18:11:28 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 223.672 uVs 18:11:31 I2Y Average and RMS of Measurements VT = -214.807 +- 14.430 uVs I2Y = -19.762 +- 1.328 uTm^2 18:11:35 Wire Position: X0 = 2.0000 mm, Y0 = 0.0000 mm 18:11:40 Integrated Voltage Measurement... 18:11:49 Voltage Offset Correction: v_off_ini = 2150.179 uV, v_off_fin = 2158.047 uV 18:11:49 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -83.934 uVs 18:11:49 Integrated Voltage Measurement... 18:11:58 Voltage Offset Correction: v_off_ini = 2163.764 uV, v_off_fin = 2156.917 uV 18:11:58 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 105.432 uVs 18:11:58 Integrated Voltage Measurement... 18:12:07 Voltage Offset Correction: v_off_ini = 2155.902 uV, v_off_fin = 2160.519 uV 18:12:07 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -76.732 uVs 18:12:07 Integrated Voltage Measurement... 18:12:16 Voltage Offset Correction: v_off_ini = 2165.407 uV, v_off_fin = 2165.565 uV 18:12:16 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 76.358 uVs 18:12:16 Integrated Voltage Measurement... 18:12:25 Voltage Offset Correction: v_off_ini = 2151.336 uV, v_off_fin = 2154.104 uV 18:12:25 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -66.489 uVs 18:12:25 Integrated Voltage Measurement... 18:12:33 Voltage Offset Correction: v_off_ini = 2184.609 uV, v_off_fin = 2202.678 uV 18:12:34 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 74.058 uVs 18:12:38 I1X Average and RMS of Measurements VT = -80.501 +- 10.350 uVs I1X = 16.100 +- 2.070 uTm 18:12:40 Wire Position: X0 = 2.0000 mm, Y0 = 0.0000 mm 18:12:44 Integrated Voltage Measurement... 18:12:53 Voltage Offset Correction: v_off_ini = 2189.597 uV, v_off_fin = 2177.736 uV 18:12:53 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -230.430 uVs 18:12:53 Integrated Voltage Measurement... 18:13:02 Voltage Offset Correction: v_off_ini = 2180.841 uV, v_off_fin = 2190.386 uV 18:13:02 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 218.281 uVs 18:13:02 Integrated Voltage Measurement... 18:13:11 Voltage Offset Correction: v_off_ini = 2172.742 uV, v_off_fin = 2149.023 uV 18:13:11 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -204.842 uVs 18:13:11 Integrated Voltage Measurement... 18:13:20 Voltage Offset Correction: v_off_ini = 2183.202 uV, v_off_fin = 2173.157 uV 18:13:20 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 187.584 uVs 18:13:20 Integrated Voltage Measurement... 18:13:29 Voltage Offset Correction: v_off_ini = 2176.278 uV, v_off_fin = 2163.314 uV 18:13:29 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -200.739 uVs 18:13:29 Integrated Voltage Measurement... 18:13:38 Voltage Offset Correction: v_off_ini = 2180.098 uV, v_off_fin = 2192.168 uV 18:13:38 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 211.981 uVs 18:13:42 I1Y Average and RMS of Measurements VT = -208.976 +- 11.637 uVs I1Y = -41.795 +- 2.327 uTm 18:13:47 Wire Position: X0 = 2.0000 mm, Y0 = 0.0000 mm 18:13:50 Integrated Voltage Measurement... 18:13:59 Voltage Offset Correction: v_off_ini = 2132.499 uV, v_off_fin = 2151.515 uV 18:13:59 Integrated Voltage Measurement dXYus = -1.0001 mm, dXYds = 0.0000 mm dXY = -1.0001 mm VT = -62.112 uVs 18:13:59 Integrated Voltage Measurement... 18:14:08 Voltage Offset Correction: v_off_ini = 2149.914 uV, v_off_fin = 2144.698 uV 18:14:08 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 52.130 uVs 18:14:08 Integrated Voltage Measurement... 18:14:17 Voltage Offset Correction: v_off_ini = 2137.551 uV, v_off_fin = 2150.088 uV 18:14:17 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -52.580 uVs 18:14:17 Integrated Voltage Measurement... 18:14:26 Voltage Offset Correction: v_off_ini = 2180.595 uV, v_off_fin = 2197.739 uV 18:14:26 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 49.958 uVs 18:14:26 Integrated Voltage Measurement... 18:14:35 Voltage Offset Correction: v_off_ini = 2185.182 uV, v_off_fin = 2177.050 uV 18:14:35 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -52.180 uVs 18:14:35 Integrated Voltage Measurement... 18:14:44 Voltage Offset Correction: v_off_ini = 2173.926 uV, v_off_fin = 2181.265 uV 18:14:44 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 59.619 uVs 18:14:46 I2X Average and RMS of Measurements VT = -54.763 +- 2.521 uVs I2X = 5.038 +- 0.232 uTm^2 18:14:52 Wire Position: X0 = 2.0000 mm, Y0 = 0.0000 mm 18:14:54 Integrated Voltage Measurement... 18:15:03 Voltage Offset Correction: v_off_ini = 2186.712 uV, v_off_fin = 2190.229 uV 18:15:03 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -238.519 uVs 18:15:03 Integrated Voltage Measurement... 18:15:12 Voltage Offset Correction: v_off_ini = 2184.427 uV, v_off_fin = 2181.450 uV 18:15:12 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 210.554 uVs 18:15:12 Integrated Voltage Measurement... 18:15:21 Voltage Offset Correction: v_off_ini = 2172.562 uV, v_off_fin = 2166.599 uV 18:15:21 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -231.505 uVs 18:15:21 Integrated Voltage Measurement... 18:15:30 Voltage Offset Correction: v_off_ini = 2200.535 uV, v_off_fin = 2199.193 uV 18:15:30 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 257.220 uVs 18:15:30 Integrated Voltage Measurement... 18:15:39 Voltage Offset Correction: v_off_ini = 2203.404 uV, v_off_fin = 2193.228 uV 18:15:39 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -197.328 uVs 18:15:39 Integrated Voltage Measurement... 18:15:48 Voltage Offset Correction: v_off_ini = 2196.446 uV, v_off_fin = 2189.101 uV 18:15:48 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 237.092 uVs 18:15:51 I2Y Average and RMS of Measurements VT = -228.703 +- 11.470 uVs I2Y = -21.041 +- 1.055 uTm^2 18:15:58 Wire Position: X0 = 0.0000 mm, Y0 = -1.0000 mm 18:16:03 Integrated Voltage Measurement... 18:16:12 Voltage Offset Correction: v_off_ini = 2193.915 uV, v_off_fin = 2178.796 uV 18:16:12 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -124.581 uVs 18:16:12 Integrated Voltage Measurement... 18:16:20 Voltage Offset Correction: v_off_ini = 2188.706 uV, v_off_fin = 2195.020 uV 18:16:21 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 148.478 uVs 18:16:21 Integrated Voltage Measurement... 18:16:29 Voltage Offset Correction: v_off_ini = 2181.115 uV, v_off_fin = 2173.541 uV 18:16:30 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -121.618 uVs 18:16:30 Integrated Voltage Measurement... 18:16:38 Voltage Offset Correction: v_off_ini = 2199.948 uV, v_off_fin = 2214.375 uV 18:16:39 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 131.526 uVs 18:16:39 Integrated Voltage Measurement... 18:16:47 Voltage Offset Correction: v_off_ini = 2193.745 uV, v_off_fin = 2179.980 uV 18:16:48 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -139.536 uVs 18:16:48 Integrated Voltage Measurement... 18:16:56 Voltage Offset Correction: v_off_ini = 2173.873 uV, v_off_fin = 2200.061 uV 18:16:57 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 138.482 uVs 18:17:01 I1X Average and RMS of Measurements VT = -134.037 +- 5.374 uVs I1X = 26.808 +- 1.076 uTm 18:17:03 Wire Position: X0 = 0.0000 mm, Y0 = -1.0000 mm 18:17:07 Integrated Voltage Measurement... 18:17:16 Voltage Offset Correction: v_off_ini = 2217.351 uV, v_off_fin = 2203.481 uV 18:17:16 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -236.384 uVs 18:17:16 Integrated Voltage Measurement... 18:17:25 Voltage Offset Correction: v_off_ini = 2190.112 uV, v_off_fin = 2203.687 uV 18:17:25 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 229.937 uVs 18:17:25 Integrated Voltage Measurement... 18:17:34 Voltage Offset Correction: v_off_ini = 2185.749 uV, v_off_fin = 2163.557 uV 18:17:34 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -245.693 uVs 18:17:34 Integrated Voltage Measurement... 18:17:43 Voltage Offset Correction: v_off_ini = 2194.209 uV, v_off_fin = 2203.425 uV 18:17:43 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 229.350 uVs 18:17:43 Integrated Voltage Measurement... 18:17:52 Voltage Offset Correction: v_off_ini = 2167.715 uV, v_off_fin = 2161.815 uV 18:17:52 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -236.072 uVs 18:17:52 Integrated Voltage Measurement... 18:18:01 Voltage Offset Correction: v_off_ini = 2172.777 uV, v_off_fin = 2209.328 uV 18:18:01 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 229.333 uVs 18:18:05 I1Y Average and RMS of Measurements VT = -234.462 +- 2.172 uVs I1Y = -46.892 +- 0.434 uTm 18:18:08 Wire Position: X0 = 0.0000 mm, Y0 = -1.0000 mm 18:18:11 Integrated Voltage Measurement... 18:18:20 Voltage Offset Correction: v_off_ini = 2207.902 uV, v_off_fin = 2188.648 uV 18:18:20 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -125.050 uVs 18:18:20 Integrated Voltage Measurement... 18:18:29 Voltage Offset Correction: v_off_ini = 2220.010 uV, v_off_fin = 2226.488 uV 18:18:29 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 109.026 uVs 18:18:29 Integrated Voltage Measurement... 18:18:38 Voltage Offset Correction: v_off_ini = 2217.450 uV, v_off_fin = 2194.685 uV 18:18:38 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -103.626 uVs 18:18:38 Integrated Voltage Measurement... 18:18:47 Voltage Offset Correction: v_off_ini = 2193.104 uV, v_off_fin = 2210.284 uV 18:18:47 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 114.404 uVs 18:18:47 Integrated Voltage Measurement... 18:18:56 Voltage Offset Correction: v_off_ini = 2211.750 uV, v_off_fin = 2210.219 uV 18:18:56 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -112.687 uVs 18:18:56 Integrated Voltage Measurement... 18:19:05 Voltage Offset Correction: v_off_ini = 2208.351 uV, v_off_fin = 2215.226 uV 18:19:05 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 97.684 uVs 18:19:07 I2X Average and RMS of Measurements VT = -110.413 +- 4.939 uVs I2X = 10.158 +- 0.454 uTm^2 18:19:09 Wire Position: X0 = 0.0000 mm, Y0 = -1.0000 mm 18:19:12 Integrated Voltage Measurement... 18:19:21 Voltage Offset Correction: v_off_ini = 2204.904 uV, v_off_fin = 2216.257 uV 18:19:21 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -220.265 uVs 18:19:21 Integrated Voltage Measurement... 18:19:30 Voltage Offset Correction: v_off_ini = 2194.700 uV, v_off_fin = 2206.495 uV 18:19:30 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 237.219 uVs 18:19:30 Integrated Voltage Measurement... 18:19:39 Voltage Offset Correction: v_off_ini = 2204.091 uV, v_off_fin = 2210.068 uV 18:19:39 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -214.932 uVs 18:19:39 Integrated Voltage Measurement... 18:19:48 Voltage Offset Correction: v_off_ini = 2212.276 uV, v_off_fin = 2201.946 uV 18:19:48 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 225.652 uVs 18:19:48 Integrated Voltage Measurement... 18:19:57 Voltage Offset Correction: v_off_ini = 2195.993 uV, v_off_fin = 2203.291 uV 18:19:57 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -220.987 uVs 18:19:57 Integrated Voltage Measurement... 18:20:06 Voltage Offset Correction: v_off_ini = 2218.501 uV, v_off_fin = 2220.342 uV 18:20:06 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 243.805 uVs 18:20:08 I2Y Average and RMS of Measurements VT = -227.144 +- 5.069 uVs I2Y = -20.897 +- 0.466 uTm^2 18:20:13 Wire Position: X0 = 0.0000 mm, Y0 = -0.5000 mm 18:20:17 Integrated Voltage Measurement... 18:20:26 Voltage Offset Correction: v_off_ini = 2214.501 uV, v_off_fin = 2201.403 uV 18:20:26 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -115.635 uVs 18:20:26 Integrated Voltage Measurement... 18:20:35 Voltage Offset Correction: v_off_ini = 2199.810 uV, v_off_fin = 2207.069 uV 18:20:35 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = 127.472 uVs 18:20:35 Integrated Voltage Measurement... 18:20:44 Voltage Offset Correction: v_off_ini = 2197.821 uV, v_off_fin = 2183.817 uV 18:20:44 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -116.724 uVs 18:20:44 Integrated Voltage Measurement... 18:20:53 Voltage Offset Correction: v_off_ini = 2192.067 uV, v_off_fin = 2201.054 uV 18:20:53 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 118.429 uVs 18:20:53 Integrated Voltage Measurement... 18:21:02 Voltage Offset Correction: v_off_ini = 2197.536 uV, v_off_fin = 2195.158 uV 18:21:02 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -122.146 uVs 18:21:02 Integrated Voltage Measurement... 18:21:11 Voltage Offset Correction: v_off_ini = 2197.419 uV, v_off_fin = 2191.666 uV 18:21:11 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 108.923 uVs 18:21:15 I1X Average and RMS of Measurements VT = -118.221 +- 2.499 uVs I1X = 23.645 +- 0.500 uTm 18:21:17 Wire Position: X0 = 0.0000 mm, Y0 = -0.5000 mm 18:21:22 Integrated Voltage Measurement... 18:21:31 Voltage Offset Correction: v_off_ini = 2191.196 uV, v_off_fin = 2169.699 uV 18:21:31 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -231.660 uVs 18:21:31 Integrated Voltage Measurement... 18:21:40 Voltage Offset Correction: v_off_ini = 2168.898 uV, v_off_fin = 2180.238 uV 18:21:40 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 224.440 uVs 18:21:40 Integrated Voltage Measurement... 18:21:48 Voltage Offset Correction: v_off_ini = 2166.642 uV, v_off_fin = 2168.016 uV 18:21:49 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -216.227 uVs 18:21:49 Integrated Voltage Measurement... 18:21:57 Voltage Offset Correction: v_off_ini = 2185.657 uV, v_off_fin = 2179.971 uV 18:21:57 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 221.984 uVs 18:21:58 Integrated Voltage Measurement... 18:22:06 Voltage Offset Correction: v_off_ini = 2182.878 uV, v_off_fin = 2173.561 uV 18:22:06 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -223.364 uVs 18:22:06 Integrated Voltage Measurement... 18:22:15 Voltage Offset Correction: v_off_ini = 2183.309 uV, v_off_fin = 2181.741 uV 18:22:15 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 207.300 uVs 18:22:20 I1Y Average and RMS of Measurements VT = -220.829 +- 5.333 uVs I1Y = -44.166 +- 1.067 uTm 18:22:22 Wire Position: X0 = 0.0000 mm, Y0 = -0.5000 mm 18:22:25 Integrated Voltage Measurement... 18:22:34 Voltage Offset Correction: v_off_ini = 2181.709 uV, v_off_fin = 2154.094 uV 18:22:34 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -81.180 uVs 18:22:34 Integrated Voltage Measurement... 18:22:43 Voltage Offset Correction: v_off_ini = 2153.255 uV, v_off_fin = 2151.457 uV 18:22:43 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 129.673 uVs 18:22:43 Integrated Voltage Measurement... 18:22:52 Voltage Offset Correction: v_off_ini = 2146.103 uV, v_off_fin = 2165.368 uV 18:22:52 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -130.867 uVs 18:22:52 Integrated Voltage Measurement... 18:23:01 Voltage Offset Correction: v_off_ini = 2188.159 uV, v_off_fin = 2193.817 uV 18:23:01 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 99.508 uVs 18:23:01 Integrated Voltage Measurement... 18:23:10 Voltage Offset Correction: v_off_ini = 2158.591 uV, v_off_fin = 2146.163 uV 18:23:10 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -130.795 uVs 18:23:10 Integrated Voltage Measurement... 18:23:19 Voltage Offset Correction: v_off_ini = 2149.280 uV, v_off_fin = 2162.531 uV 18:23:19 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 86.256 uVs 18:23:21 I2X Average and RMS of Measurements VT = -109.713 +- 4.072 uVs I2X = 10.094 +- 0.375 uTm^2 18:23:24 Wire Position: X0 = 0.0000 mm, Y0 = -0.5000 mm 18:23:26 Integrated Voltage Measurement... 18:23:35 Voltage Offset Correction: v_off_ini = 2155.166 uV, v_off_fin = 2162.524 uV 18:23:35 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -228.490 uVs 18:23:35 Integrated Voltage Measurement... 18:23:44 Voltage Offset Correction: v_off_ini = 2163.669 uV, v_off_fin = 2151.887 uV 18:23:44 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 228.722 uVs 18:23:44 Integrated Voltage Measurement... 18:23:53 Voltage Offset Correction: v_off_ini = 2149.098 uV, v_off_fin = 2168.181 uV 18:23:53 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -227.699 uVs 18:23:53 Integrated Voltage Measurement... 18:24:02 Voltage Offset Correction: v_off_ini = 2156.410 uV, v_off_fin = 2144.687 uV 18:24:02 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 219.289 uVs 18:24:02 Integrated Voltage Measurement... 18:24:11 Voltage Offset Correction: v_off_ini = 2143.115 uV, v_off_fin = 2151.780 uV 18:24:11 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -201.396 uVs 18:24:11 Integrated Voltage Measurement... 18:24:20 Voltage Offset Correction: v_off_ini = 2167.759 uV, v_off_fin = 2169.671 uV 18:24:20 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 211.844 uVs 18:24:23 I2Y Average and RMS of Measurements VT = -219.573 +- 9.394 uVs I2Y = -20.201 +- 0.864 uTm^2 18:24:27 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 18:24:31 Integrated Voltage Measurement... 18:24:40 Voltage Offset Correction: v_off_ini = 2167.021 uV, v_off_fin = 2158.645 uV 18:24:40 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -122.111 uVs 18:24:40 Integrated Voltage Measurement... 18:24:49 Voltage Offset Correction: v_off_ini = 2154.863 uV, v_off_fin = 2141.391 uV 18:24:49 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 105.748 uVs 18:24:49 Integrated Voltage Measurement... 18:24:58 Voltage Offset Correction: v_off_ini = 2154.034 uV, v_off_fin = 2146.123 uV 18:24:58 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -126.464 uVs 18:24:58 Integrated Voltage Measurement... 18:25:07 Voltage Offset Correction: v_off_ini = 2155.451 uV, v_off_fin = 2169.692 uV 18:25:07 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 111.888 uVs 18:25:07 Integrated Voltage Measurement... 18:25:16 Voltage Offset Correction: v_off_ini = 2162.735 uV, v_off_fin = 2182.276 uV 18:25:16 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -116.356 uVs 18:25:16 Integrated Voltage Measurement... 18:25:25 Voltage Offset Correction: v_off_ini = 2179.944 uV, v_off_fin = 2191.511 uV 18:25:25 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 111.981 uVs 18:25:29 I1X Average and RMS of Measurements VT = -115.758 +- 2.419 uVs I1X = 23.152 +- 0.484 uTm 18:25:32 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 18:25:36 Integrated Voltage Measurement... 18:25:45 Voltage Offset Correction: v_off_ini = 2178.314 uV, v_off_fin = 2159.107 uV 18:25:45 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -208.489 uVs 18:25:45 Integrated Voltage Measurement... 18:25:54 Voltage Offset Correction: v_off_ini = 2193.696 uV, v_off_fin = 2202.721 uV 18:25:54 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 201.493 uVs 18:25:54 Integrated Voltage Measurement... 18:26:03 Voltage Offset Correction: v_off_ini = 2161.016 uV, v_off_fin = 2170.007 uV 18:26:03 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -185.663 uVs 18:26:03 Integrated Voltage Measurement... 18:26:12 Voltage Offset Correction: v_off_ini = 2172.810 uV, v_off_fin = 2184.991 uV 18:26:12 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 182.518 uVs 18:26:12 Integrated Voltage Measurement... 18:26:21 Voltage Offset Correction: v_off_ini = 2201.476 uV, v_off_fin = 2175.002 uV 18:26:21 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -207.128 uVs 18:26:21 Integrated Voltage Measurement... 18:26:30 Voltage Offset Correction: v_off_ini = 2183.539 uV, v_off_fin = 2190.390 uV 18:26:30 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 193.729 uVs 18:26:34 I1Y Average and RMS of Measurements VT = -196.503 +- 8.973 uVs I1Y = -39.301 +- 1.795 uTm 18:26:36 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 18:26:39 Integrated Voltage Measurement... 18:26:48 Voltage Offset Correction: v_off_ini = 2209.215 uV, v_off_fin = 2202.810 uV 18:26:48 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -102.041 uVs 18:26:48 Integrated Voltage Measurement... 18:26:57 Voltage Offset Correction: v_off_ini = 2199.075 uV, v_off_fin = 2208.783 uV 18:26:57 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 86.315 uVs 18:26:57 Integrated Voltage Measurement... 18:27:06 Voltage Offset Correction: v_off_ini = 2196.061 uV, v_off_fin = 2188.006 uV 18:27:06 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -116.443 uVs 18:27:06 Integrated Voltage Measurement... 18:27:15 Voltage Offset Correction: v_off_ini = 2207.815 uV, v_off_fin = 2205.776 uV 18:27:15 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 108.067 uVs 18:27:15 Integrated Voltage Measurement... 18:27:24 Voltage Offset Correction: v_off_ini = 2209.355 uV, v_off_fin = 2201.256 uV 18:27:24 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -119.935 uVs 18:27:24 Integrated Voltage Measurement... 18:27:33 Voltage Offset Correction: v_off_ini = 2204.384 uV, v_off_fin = 2214.254 uV 18:27:33 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 88.825 uVs 18:27:35 I2X Average and RMS of Measurements VT = -103.604 +- 7.400 uVs I2X = 9.532 +- 0.681 uTm^2 18:27:38 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 18:27:40 Integrated Voltage Measurement... 18:27:49 Voltage Offset Correction: v_off_ini = 2201.851 uV, v_off_fin = 2195.974 uV 18:27:49 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -202.908 uVs 18:27:49 Integrated Voltage Measurement... 18:27:58 Voltage Offset Correction: v_off_ini = 2192.174 uV, v_off_fin = 2203.278 uV 18:27:58 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 206.926 uVs 18:27:58 Integrated Voltage Measurement... 18:28:07 Voltage Offset Correction: v_off_ini = 2208.657 uV, v_off_fin = 2192.284 uV 18:28:07 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -202.848 uVs 18:28:08 Integrated Voltage Measurement... 18:28:16 Voltage Offset Correction: v_off_ini = 2193.812 uV, v_off_fin = 2204.720 uV 18:28:17 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 220.162 uVs 18:28:17 Integrated Voltage Measurement... 18:28:26 Voltage Offset Correction: v_off_ini = 2199.696 uV, v_off_fin = 2206.033 uV 18:28:26 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -188.263 uVs 18:28:26 Integrated Voltage Measurement... 18:28:35 Voltage Offset Correction: v_off_ini = 2207.908 uV, v_off_fin = 2196.645 uV 18:28:35 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 204.027 uVs 18:28:37 I2Y Average and RMS of Measurements VT = -204.189 +- 6.292 uVs I2Y = -18.785 +- 0.579 uTm^2 18:28:41 Wire Position: X0 = 0.0000 mm, Y0 = 0.5000 mm 18:28:45 Integrated Voltage Measurement... 18:28:54 Voltage Offset Correction: v_off_ini = 2189.149 uV, v_off_fin = 2169.050 uV 18:28:54 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -108.687 uVs 18:28:54 Integrated Voltage Measurement... 18:29:03 Voltage Offset Correction: v_off_ini = 2182.041 uV, v_off_fin = 2179.336 uV 18:29:03 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 98.818 uVs 18:29:03 Integrated Voltage Measurement... 18:29:12 Voltage Offset Correction: v_off_ini = 2175.323 uV, v_off_fin = 2180.458 uV 18:29:12 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -121.848 uVs 18:29:12 Integrated Voltage Measurement... 18:29:21 Voltage Offset Correction: v_off_ini = 2195.125 uV, v_off_fin = 2196.817 uV 18:29:21 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 112.408 uVs 18:29:21 Integrated Voltage Measurement... 18:29:30 Voltage Offset Correction: v_off_ini = 2196.842 uV, v_off_fin = 2180.210 uV 18:29:30 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -117.220 uVs 18:29:30 Integrated Voltage Measurement... 18:29:39 Voltage Offset Correction: v_off_ini = 2188.107 uV, v_off_fin = 2181.354 uV 18:29:39 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = 109.422 uVs 18:29:43 I1X Average and RMS of Measurements VT = -111.401 +- 5.627 uVs I1X = 22.280 +- 1.125 uTm 18:29:46 Wire Position: X0 = 0.0000 mm, Y0 = 0.5000 mm 18:29:50 Integrated Voltage Measurement... 18:29:59 Voltage Offset Correction: v_off_ini = 2237.558 uV, v_off_fin = 2213.442 uV 18:29:59 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -175.366 uVs 18:29:59 Integrated Voltage Measurement... 18:30:08 Voltage Offset Correction: v_off_ini = 2244.114 uV, v_off_fin = 2228.764 uV 18:30:08 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 207.008 uVs 18:30:08 Integrated Voltage Measurement... 18:30:17 Voltage Offset Correction: v_off_ini = 2226.457 uV, v_off_fin = 2214.843 uV 18:30:17 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -190.134 uVs 18:30:17 Integrated Voltage Measurement... 18:30:26 Voltage Offset Correction: v_off_ini = 2210.364 uV, v_off_fin = 2231.029 uV 18:30:26 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 200.600 uVs 18:30:26 Integrated Voltage Measurement... 18:30:35 Voltage Offset Correction: v_off_ini = 2216.865 uV, v_off_fin = 2208.575 uV 18:30:35 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -170.036 uVs 18:30:35 Integrated Voltage Measurement... 18:30:44 Voltage Offset Correction: v_off_ini = 2242.139 uV, v_off_fin = 2239.056 uV 18:30:44 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 202.601 uVs 18:30:48 I1Y Average and RMS of Measurements VT = -190.958 +- 3.697 uVs I1Y = -38.190 +- 0.739 uTm 18:30:51 Wire Position: X0 = 0.0000 mm, Y0 = 0.5000 mm 18:30:53 Integrated Voltage Measurement... 18:31:02 Voltage Offset Correction: v_off_ini = 2191.702 uV, v_off_fin = 2200.232 uV 18:31:02 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -116.116 uVs 18:31:02 Integrated Voltage Measurement... 18:31:11 Voltage Offset Correction: v_off_ini = 2219.989 uV, v_off_fin = 2229.877 uV 18:31:11 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 93.596 uVs 18:31:11 Integrated Voltage Measurement... 18:31:20 Voltage Offset Correction: v_off_ini = 2240.113 uV, v_off_fin = 2207.413 uV 18:31:20 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -88.008 uVs 18:31:20 Integrated Voltage Measurement... 18:31:29 Voltage Offset Correction: v_off_ini = 2214.354 uV, v_off_fin = 2236.511 uV 18:31:29 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 112.317 uVs 18:31:29 Integrated Voltage Measurement... 18:31:38 Voltage Offset Correction: v_off_ini = 2235.693 uV, v_off_fin = 2221.342 uV 18:31:38 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -84.757 uVs 18:31:38 Integrated Voltage Measurement... 18:31:47 Voltage Offset Correction: v_off_ini = 2223.471 uV, v_off_fin = 2233.309 uV 18:31:47 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 85.433 uVs 18:31:50 I2X Average and RMS of Measurements VT = -96.704 +- 8.430 uVs I2X = 8.897 +- 0.776 uTm^2 18:31:52 Wire Position: X0 = 0.0000 mm, Y0 = 0.5000 mm 18:31:55 Integrated Voltage Measurement... 18:32:04 Voltage Offset Correction: v_off_ini = 2202.591 uV, v_off_fin = 2185.376 uV 18:32:04 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -185.096 uVs 18:32:04 Integrated Voltage Measurement... 18:32:13 Voltage Offset Correction: v_off_ini = 2186.648 uV, v_off_fin = 2193.009 uV 18:32:13 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 180.017 uVs 18:32:13 Integrated Voltage Measurement... 18:32:22 Voltage Offset Correction: v_off_ini = 2199.113 uV, v_off_fin = 2207.056 uV 18:32:22 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -185.123 uVs 18:32:22 Integrated Voltage Measurement... 18:32:31 Voltage Offset Correction: v_off_ini = 2203.137 uV, v_off_fin = 2195.813 uV 18:32:31 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 187.635 uVs 18:32:31 Integrated Voltage Measurement... 18:32:40 Voltage Offset Correction: v_off_ini = 2202.636 uV, v_off_fin = 2219.974 uV 18:32:40 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -191.755 uVs 18:32:40 Integrated Voltage Measurement... 18:32:49 Voltage Offset Correction: v_off_ini = 2231.920 uV, v_off_fin = 2213.342 uV 18:32:49 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 199.361 uVs 18:32:51 I2Y Average and RMS of Measurements VT = -188.164 +- 5.456 uVs I2Y = -17.311 +- 0.502 uTm^2 18:32:55 Wire Position: X0 = 0.0000 mm, Y0 = 1.0000 mm 18:32:59 Integrated Voltage Measurement... 18:33:08 Voltage Offset Correction: v_off_ini = 2202.680 uV, v_off_fin = 2198.567 uV 18:33:08 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -98.345 uVs 18:33:08 Integrated Voltage Measurement... 18:33:17 Voltage Offset Correction: v_off_ini = 2198.555 uV, v_off_fin = 2219.278 uV 18:33:17 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 89.257 uVs 18:33:17 Integrated Voltage Measurement... 18:33:26 Voltage Offset Correction: v_off_ini = 2197.177 uV, v_off_fin = 2201.573 uV 18:33:26 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -120.370 uVs 18:33:26 Integrated Voltage Measurement... 18:33:35 Voltage Offset Correction: v_off_ini = 2198.613 uV, v_off_fin = 2200.538 uV 18:33:35 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 110.579 uVs 18:33:35 Integrated Voltage Measurement... 18:33:44 Voltage Offset Correction: v_off_ini = 2197.585 uV, v_off_fin = 2213.703 uV 18:33:44 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -124.166 uVs 18:33:44 Integrated Voltage Measurement... 18:33:53 Voltage Offset Correction: v_off_ini = 2217.477 uV, v_off_fin = 2207.675 uV 18:33:53 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 122.839 uVs 18:33:57 I1X Average and RMS of Measurements VT = -110.926 +- 12.545 uVs I1X = 22.186 +- 2.509 uTm 18:34:00 Wire Position: X0 = 0.0000 mm, Y0 = 1.0000 mm 18:34:04 Integrated Voltage Measurement... 18:34:13 Voltage Offset Correction: v_off_ini = 2197.387 uV, v_off_fin = 2199.215 uV 18:34:13 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -174.640 uVs 18:34:13 Integrated Voltage Measurement... 18:34:22 Voltage Offset Correction: v_off_ini = 2215.960 uV, v_off_fin = 2207.162 uV 18:34:22 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 192.280 uVs 18:34:22 Integrated Voltage Measurement... 18:34:31 Voltage Offset Correction: v_off_ini = 2194.234 uV, v_off_fin = 2194.215 uV 18:34:31 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -173.499 uVs 18:34:31 Integrated Voltage Measurement... 18:34:40 Voltage Offset Correction: v_off_ini = 2194.864 uV, v_off_fin = 2215.479 uV 18:34:40 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 177.852 uVs 18:34:40 Integrated Voltage Measurement... 18:34:49 Voltage Offset Correction: v_off_ini = 2219.313 uV, v_off_fin = 2192.127 uV 18:34:49 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -177.111 uVs 18:34:49 Integrated Voltage Measurement... 18:34:58 Voltage Offset Correction: v_off_ini = 2202.535 uV, v_off_fin = 2208.681 uV 18:34:58 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 177.835 uVs 18:35:02 I1Y Average and RMS of Measurements VT = -178.869 +- 3.328 uVs I1Y = -35.773 +- 0.665 uTm 18:35:05 Wire Position: X0 = 0.0000 mm, Y0 = 1.0000 mm 18:35:07 Integrated Voltage Measurement... 18:35:16 Voltage Offset Correction: v_off_ini = 2217.861 uV, v_off_fin = 2192.667 uV 18:35:16 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -99.195 uVs 18:35:16 Integrated Voltage Measurement... 18:35:25 Voltage Offset Correction: v_off_ini = 2200.748 uV, v_off_fin = 2207.671 uV 18:35:25 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 83.700 uVs 18:35:25 Integrated Voltage Measurement... 18:35:34 Voltage Offset Correction: v_off_ini = 2184.136 uV, v_off_fin = 2167.984 uV 18:35:34 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -102.680 uVs 18:35:34 Integrated Voltage Measurement... 18:35:43 Voltage Offset Correction: v_off_ini = 2187.911 uV, v_off_fin = 2172.420 uV 18:35:43 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 99.521 uVs 18:35:43 Integrated Voltage Measurement... 18:35:52 Voltage Offset Correction: v_off_ini = 2168.746 uV, v_off_fin = 2166.940 uV 18:35:52 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -93.184 uVs 18:35:52 Integrated Voltage Measurement... 18:36:01 Voltage Offset Correction: v_off_ini = 2190.068 uV, v_off_fin = 2202.980 uV 18:36:01 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 95.647 uVs 18:36:04 I2X Average and RMS of Measurements VT = -95.655 +- 4.037 uVs I2X = 8.800 +- 0.371 uTm^2 18:36:06 Wire Position: X0 = 0.0000 mm, Y0 = 1.0000 mm 18:36:09 Integrated Voltage Measurement... 18:36:18 Voltage Offset Correction: v_off_ini = 2202.371 uV, v_off_fin = 2220.446 uV 18:36:18 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -174.428 uVs 18:36:18 Integrated Voltage Measurement... 18:36:27 Voltage Offset Correction: v_off_ini = 2208.985 uV, v_off_fin = 2213.960 uV 18:36:27 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 169.116 uVs 18:36:27 Integrated Voltage Measurement... 18:36:36 Voltage Offset Correction: v_off_ini = 2209.309 uV, v_off_fin = 2211.128 uV 18:36:36 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -172.434 uVs 18:36:36 Integrated Voltage Measurement... 18:36:45 Voltage Offset Correction: v_off_ini = 2216.054 uV, v_off_fin = 2218.290 uV 18:36:45 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 188.403 uVs 18:36:45 Integrated Voltage Measurement... 18:36:54 Voltage Offset Correction: v_off_ini = 2212.850 uV, v_off_fin = 2216.601 uV 18:36:54 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -158.950 uVs 18:36:54 Integrated Voltage Measurement... 18:37:03 Voltage Offset Correction: v_off_ini = 2217.017 uV, v_off_fin = 2217.085 uV 18:37:03 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 177.509 uVs 18:37:05 I2Y Average and RMS of Measurements VT = -173.473 +- 5.119 uVs I2Y = -15.960 +- 0.471 uTm^2 18:37:22 Wire Position: X0 = -2.0000 mm, Y0 = 0.0000 mm 18:37:27 Integrated Voltage Measurement... 18:37:35 Voltage Offset Correction: v_off_ini = 2195.931 uV, v_off_fin = 2194.058 uV 18:37:35 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -165.554 uVs 18:37:36 Integrated Voltage Measurement... 18:37:44 Voltage Offset Correction: v_off_ini = 2213.063 uV, v_off_fin = 2220.395 uV 18:37:44 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 194.391 uVs 18:37:45 Integrated Voltage Measurement... 18:37:53 Voltage Offset Correction: v_off_ini = 2192.175 uV, v_off_fin = 2210.120 uV 18:37:53 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -145.980 uVs 18:37:54 Integrated Voltage Measurement... 18:38:02 Voltage Offset Correction: v_off_ini = 2222.630 uV, v_off_fin = 2219.797 uV 18:38:02 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 169.334 uVs 18:38:03 Integrated Voltage Measurement... 18:38:11 Voltage Offset Correction: v_off_ini = 2221.821 uV, v_off_fin = 2215.640 uV 18:38:11 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -146.448 uVs 18:38:12 Integrated Voltage Measurement... 18:38:20 Voltage Offset Correction: v_off_ini = 2223.387 uV, v_off_fin = 2225.740 uV 18:38:20 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 168.679 uVs 18:38:25 I1X Average and RMS of Measurements VT = -165.064 +- 10.542 uVs I1X = 33.013 +- 2.108 uTm 18:38:27 Wire Position: X0 = -2.0000 mm, Y0 = 0.0000 mm 18:38:31 Integrated Voltage Measurement... 18:38:40 Voltage Offset Correction: v_off_ini = 2229.090 uV, v_off_fin = 2228.834 uV 18:38:40 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -153.242 uVs 18:38:40 Integrated Voltage Measurement... 18:38:49 Voltage Offset Correction: v_off_ini = 2221.231 uV, v_off_fin = 2237.156 uV 18:38:49 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 163.752 uVs 18:38:49 Integrated Voltage Measurement... 18:38:58 Voltage Offset Correction: v_off_ini = 2220.858 uV, v_off_fin = 2208.054 uV 18:38:58 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -170.292 uVs 18:38:58 Integrated Voltage Measurement... 18:39:07 Voltage Offset Correction: v_off_ini = 2217.066 uV, v_off_fin = 2229.597 uV 18:39:07 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 172.314 uVs 18:39:07 Integrated Voltage Measurement... 18:39:16 Voltage Offset Correction: v_off_ini = 2219.907 uV, v_off_fin = 2219.703 uV 18:39:16 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -181.848 uVs 18:39:16 Integrated Voltage Measurement... 18:39:25 Voltage Offset Correction: v_off_ini = 2244.178 uV, v_off_fin = 2221.480 uV 18:39:25 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 167.718 uVs 18:39:29 I1Y Average and RMS of Measurements VT = -168.194 +- 7.003 uVs I1Y = -33.639 +- 1.401 uTm 18:39:35 Wire Position: X0 = -2.0000 mm, Y0 = 0.0000 mm 18:39:37 Integrated Voltage Measurement... 18:39:46 Voltage Offset Correction: v_off_ini = 2243.040 uV, v_off_fin = 2240.644 uV 18:39:46 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -150.872 uVs 18:39:46 Integrated Voltage Measurement... 18:39:55 Voltage Offset Correction: v_off_ini = 2249.156 uV, v_off_fin = 2267.145 uV 18:39:55 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 152.630 uVs 18:39:55 Integrated Voltage Measurement... 18:40:04 Voltage Offset Correction: v_off_ini = 2256.916 uV, v_off_fin = 2234.349 uV 18:40:04 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -141.542 uVs 18:40:04 Integrated Voltage Measurement... 18:40:13 Voltage Offset Correction: v_off_ini = 2244.626 uV, v_off_fin = 2271.672 uV 18:40:13 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 148.912 uVs 18:40:13 Integrated Voltage Measurement... 18:40:22 Voltage Offset Correction: v_off_ini = 2229.931 uV, v_off_fin = 2238.054 uV 18:40:22 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -146.929 uVs 18:40:22 Integrated Voltage Measurement... 18:40:31 Voltage Offset Correction: v_off_ini = 2235.866 uV, v_off_fin = 2259.656 uV 18:40:31 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 154.529 uVs 18:40:34 I2X Average and RMS of Measurements VT = -149.236 +- 2.865 uVs I2X = 13.730 +- 0.264 uTm^2 18:40:39 Wire Position: X0 = -2.0000 mm, Y0 = 0.0000 mm 18:40:41 Integrated Voltage Measurement... 18:40:50 Voltage Offset Correction: v_off_ini = 2233.788 uV, v_off_fin = 2225.142 uV 18:40:50 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -164.210 uVs 18:40:50 Integrated Voltage Measurement... 18:40:59 Voltage Offset Correction: v_off_ini = 2227.783 uV, v_off_fin = 2220.556 uV 18:40:59 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 156.027 uVs 18:40:59 Integrated Voltage Measurement... 18:41:08 Voltage Offset Correction: v_off_ini = 2229.451 uV, v_off_fin = 2231.931 uV 18:41:08 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -172.462 uVs 18:41:08 Integrated Voltage Measurement... 18:41:17 Voltage Offset Correction: v_off_ini = 2245.041 uV, v_off_fin = 2242.779 uV 18:41:17 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 177.594 uVs 18:41:17 Integrated Voltage Measurement... 18:41:26 Voltage Offset Correction: v_off_ini = 2252.677 uV, v_off_fin = 2241.362 uV 18:41:26 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -185.690 uVs 18:41:26 Integrated Voltage Measurement... 18:41:35 Voltage Offset Correction: v_off_ini = 2241.109 uV, v_off_fin = 2246.210 uV 18:41:35 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 174.109 uVs 18:41:38 I2Y Average and RMS of Measurements VT = -171.682 +- 8.415 uVs I2Y = -15.795 +- 0.774 uTm^2 18:41:42 Wire Position: X0 = -1.0000 mm, Y0 = 0.0000 mm 18:41:47 Integrated Voltage Measurement... 18:41:55 Voltage Offset Correction: v_off_ini = 2261.570 uV, v_off_fin = 2226.247 uV 18:41:55 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -146.566 uVs 18:41:56 Integrated Voltage Measurement... 18:42:04 Voltage Offset Correction: v_off_ini = 2234.371 uV, v_off_fin = 2221.229 uV 18:42:04 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 171.779 uVs 18:42:05 Integrated Voltage Measurement... 18:42:13 Voltage Offset Correction: v_off_ini = 2240.627 uV, v_off_fin = 2217.083 uV 18:42:13 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -157.494 uVs 18:42:14 Integrated Voltage Measurement... 18:42:22 Voltage Offset Correction: v_off_ini = 2221.706 uV, v_off_fin = 2225.111 uV 18:42:22 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 165.573 uVs 18:42:23 Integrated Voltage Measurement... 18:42:31 Voltage Offset Correction: v_off_ini = 2235.371 uV, v_off_fin = 2227.104 uV 18:42:31 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -156.731 uVs 18:42:31 Integrated Voltage Measurement... 18:42:40 Voltage Offset Correction: v_off_ini = 2221.764 uV, v_off_fin = 2221.542 uV 18:42:40 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 130.542 uVs 18:42:45 I1X Average and RMS of Measurements VT = -154.781 +- 7.939 uVs I1X = 30.956 +- 1.588 uTm 18:42:47 Wire Position: X0 = -1.0000 mm, Y0 = 0.0000 mm 18:42:51 Integrated Voltage Measurement... 18:43:00 Voltage Offset Correction: v_off_ini = 2229.066 uV, v_off_fin = 2208.883 uV 18:43:00 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -185.565 uVs 18:43:00 Integrated Voltage Measurement... 18:43:09 Voltage Offset Correction: v_off_ini = 2212.106 uV, v_off_fin = 2206.526 uV 18:43:09 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 182.481 uVs 18:43:09 Integrated Voltage Measurement... 18:43:18 Voltage Offset Correction: v_off_ini = 2221.082 uV, v_off_fin = 2205.869 uV 18:43:18 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -163.207 uVs 18:43:18 Integrated Voltage Measurement... 18:43:27 Voltage Offset Correction: v_off_ini = 2211.215 uV, v_off_fin = 2205.029 uV 18:43:27 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 168.489 uVs 18:43:27 Integrated Voltage Measurement... 18:43:36 Voltage Offset Correction: v_off_ini = 2187.183 uV, v_off_fin = 2205.489 uV 18:43:36 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -162.520 uVs 18:43:36 Integrated Voltage Measurement... 18:43:45 Voltage Offset Correction: v_off_ini = 2230.328 uV, v_off_fin = 2238.296 uV 18:43:45 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 166.136 uVs 18:43:49 I1Y Average and RMS of Measurements VT = -171.400 +- 8.948 uVs I1Y = -34.280 +- 1.790 uTm 18:43:54 Wire Position: X0 = -1.0000 mm, Y0 = 0.0000 mm 18:43:57 Integrated Voltage Measurement... 18:44:06 Voltage Offset Correction: v_off_ini = 2205.188 uV, v_off_fin = 2203.951 uV 18:44:06 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -115.327 uVs 18:44:06 Integrated Voltage Measurement... 18:44:15 Voltage Offset Correction: v_off_ini = 2213.517 uV, v_off_fin = 2219.777 uV 18:44:15 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 110.675 uVs 18:44:15 Integrated Voltage Measurement... 18:44:24 Voltage Offset Correction: v_off_ini = 2218.867 uV, v_off_fin = 2218.891 uV 18:44:24 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -133.986 uVs 18:44:24 Integrated Voltage Measurement... 18:44:33 Voltage Offset Correction: v_off_ini = 2215.812 uV, v_off_fin = 2222.099 uV 18:44:33 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 115.009 uVs 18:44:33 Integrated Voltage Measurement... 18:44:42 Voltage Offset Correction: v_off_ini = 2238.353 uV, v_off_fin = 2195.416 uV 18:44:42 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -124.943 uVs 18:44:42 Integrated Voltage Measurement... 18:44:51 Voltage Offset Correction: v_off_ini = 2220.012 uV, v_off_fin = 2226.370 uV 18:44:51 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 121.558 uVs 18:44:54 I2X Average and RMS of Measurements VT = -120.250 +- 5.151 uVs I2X = 11.063 +- 0.474 uTm^2 18:44:59 Wire Position: X0 = -1.0000 mm, Y0 = 0.0000 mm 18:45:01 Integrated Voltage Measurement... 18:45:10 Voltage Offset Correction: v_off_ini = 2185.122 uV, v_off_fin = 2189.629 uV 18:45:10 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -181.182 uVs 18:45:10 Integrated Voltage Measurement... 18:45:19 Voltage Offset Correction: v_off_ini = 2187.858 uV, v_off_fin = 2196.438 uV 18:45:19 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 174.610 uVs 18:45:19 Integrated Voltage Measurement... 18:45:28 Voltage Offset Correction: v_off_ini = 2189.260 uV, v_off_fin = 2166.500 uV 18:45:28 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -156.134 uVs 18:45:28 Integrated Voltage Measurement... 18:45:37 Voltage Offset Correction: v_off_ini = 2213.918 uV, v_off_fin = 2217.063 uV 18:45:37 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 157.692 uVs 18:45:37 Integrated Voltage Measurement... 18:45:46 Voltage Offset Correction: v_off_ini = 2206.257 uV, v_off_fin = 2205.567 uV 18:45:46 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -169.719 uVs 18:45:46 Integrated Voltage Measurement... 18:45:55 Voltage Offset Correction: v_off_ini = 2213.097 uV, v_off_fin = 2192.527 uV 18:45:55 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 147.418 uVs 18:45:58 I2Y Average and RMS of Measurements VT = -164.459 +- 9.525 uVs I2Y = -15.130 +- 0.876 uTm^2 18:46:02 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 18:46:06 Integrated Voltage Measurement... 18:46:15 Voltage Offset Correction: v_off_ini = 2218.268 uV, v_off_fin = 2214.547 uV 18:46:15 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -118.107 uVs 18:46:15 Integrated Voltage Measurement... 18:46:24 Voltage Offset Correction: v_off_ini = 2221.489 uV, v_off_fin = 2207.499 uV 18:46:24 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 114.150 uVs 18:46:24 Integrated Voltage Measurement... 18:46:33 Voltage Offset Correction: v_off_ini = 2192.633 uV, v_off_fin = 2216.695 uV 18:46:33 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -108.278 uVs 18:46:33 Integrated Voltage Measurement... 18:46:42 Voltage Offset Correction: v_off_ini = 2222.567 uV, v_off_fin = 2214.087 uV 18:46:42 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 108.210 uVs 18:46:42 Integrated Voltage Measurement... 18:46:51 Voltage Offset Correction: v_off_ini = 2214.970 uV, v_off_fin = 2200.063 uV 18:46:51 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -120.082 uVs 18:46:51 Integrated Voltage Measurement... 18:47:00 Voltage Offset Correction: v_off_ini = 2227.999 uV, v_off_fin = 2220.808 uV 18:47:00 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 120.438 uVs 18:47:05 I1X Average and RMS of Measurements VT = -114.877 +- 4.985 uVs I1X = 22.975 +- 0.997 uTm 18:47:07 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 18:47:11 Integrated Voltage Measurement... 18:47:20 Voltage Offset Correction: v_off_ini = 2208.564 uV, v_off_fin = 2199.803 uV 18:47:20 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -201.000 uVs 18:47:20 Integrated Voltage Measurement... 18:47:29 Voltage Offset Correction: v_off_ini = 2212.813 uV, v_off_fin = 2211.211 uV 18:47:29 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 172.536 uVs 18:47:29 Integrated Voltage Measurement... 18:47:38 Voltage Offset Correction: v_off_ini = 2210.215 uV, v_off_fin = 2203.324 uV 18:47:38 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -172.386 uVs 18:47:38 Integrated Voltage Measurement... 18:47:47 Voltage Offset Correction: v_off_ini = 2206.594 uV, v_off_fin = 2226.131 uV 18:47:47 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 169.720 uVs 18:47:47 Integrated Voltage Measurement... 18:47:56 Voltage Offset Correction: v_off_ini = 2204.564 uV, v_off_fin = 2196.365 uV 18:47:56 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -191.309 uVs 18:47:56 Integrated Voltage Measurement... 18:48:05 Voltage Offset Correction: v_off_ini = 2193.923 uV, v_off_fin = 2195.629 uV 18:48:05 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 184.413 uVs 18:48:09 I1Y Average and RMS of Measurements VT = -181.894 +- 7.679 uVs I1Y = -36.379 +- 1.536 uTm 18:48:14 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 18:48:17 Integrated Voltage Measurement... 18:48:26 Voltage Offset Correction: v_off_ini = 2190.392 uV, v_off_fin = 2205.945 uV 18:48:26 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -93.187 uVs 18:48:26 Integrated Voltage Measurement... 18:48:35 Voltage Offset Correction: v_off_ini = 2220.337 uV, v_off_fin = 2220.125 uV 18:48:35 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 110.868 uVs 18:48:35 Integrated Voltage Measurement... 18:48:44 Voltage Offset Correction: v_off_ini = 2212.439 uV, v_off_fin = 2213.153 uV 18:48:44 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -101.225 uVs 18:48:44 Integrated Voltage Measurement... 18:48:53 Voltage Offset Correction: v_off_ini = 2215.489 uV, v_off_fin = 2240.207 uV 18:48:53 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 95.382 uVs 18:48:53 Integrated Voltage Measurement... 18:49:02 Voltage Offset Correction: v_off_ini = 2224.558 uV, v_off_fin = 2224.931 uV 18:49:02 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -111.501 uVs 18:49:02 Integrated Voltage Measurement... 18:49:11 Voltage Offset Correction: v_off_ini = 2228.341 uV, v_off_fin = 2243.128 uV 18:49:11 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 97.780 uVs 18:49:13 I2X Average and RMS of Measurements VT = -101.657 +- 2.600 uVs I2X = 9.352 +- 0.239 uTm^2 18:49:18 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 18:49:21 Integrated Voltage Measurement... 18:49:30 Voltage Offset Correction: v_off_ini = 2245.238 uV, v_off_fin = 2222.479 uV 18:49:30 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -192.451 uVs 18:49:30 Integrated Voltage Measurement... 18:49:39 Voltage Offset Correction: v_off_ini = 2241.605 uV, v_off_fin = 2237.161 uV 18:49:39 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 187.835 uVs 18:49:39 Integrated Voltage Measurement... 18:49:48 Voltage Offset Correction: v_off_ini = 2245.390 uV, v_off_fin = 2243.031 uV 18:49:48 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -181.500 uVs 18:49:48 Integrated Voltage Measurement... 18:49:57 Voltage Offset Correction: v_off_ini = 2203.836 uV, v_off_fin = 2231.858 uV 18:49:57 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 182.277 uVs 18:49:57 Integrated Voltage Measurement... 18:50:06 Voltage Offset Correction: v_off_ini = 2230.811 uV, v_off_fin = 2211.844 uV 18:50:06 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -187.322 uVs 18:50:06 Integrated Voltage Measurement... 18:50:15 Voltage Offset Correction: v_off_ini = 2218.848 uV, v_off_fin = 2217.555 uV 18:50:15 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 197.322 uVs 18:50:18 I2Y Average and RMS of Measurements VT = -188.118 +- 4.494 uVs I2Y = -17.307 +- 0.413 uTm^2 18:50:22 Wire Position: X0 = 1.0000 mm, Y0 = 0.0000 mm 18:50:26 Integrated Voltage Measurement... 18:50:35 Voltage Offset Correction: v_off_ini = 2216.486 uV, v_off_fin = 2224.159 uV 18:50:35 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -126.790 uVs 18:50:35 Integrated Voltage Measurement... 18:50:44 Voltage Offset Correction: v_off_ini = 2243.405 uV, v_off_fin = 2246.488 uV 18:50:44 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 95.956 uVs 18:50:44 Integrated Voltage Measurement... 18:50:53 Voltage Offset Correction: v_off_ini = 2239.997 uV, v_off_fin = 2235.129 uV 18:50:53 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -90.186 uVs 18:50:53 Integrated Voltage Measurement... 18:51:02 Voltage Offset Correction: v_off_ini = 2229.528 uV, v_off_fin = 2231.704 uV 18:51:02 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 108.284 uVs 18:51:02 Integrated Voltage Measurement... 18:51:11 Voltage Offset Correction: v_off_ini = 2236.135 uV, v_off_fin = 2230.646 uV 18:51:11 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -88.915 uVs 18:51:11 Integrated Voltage Measurement... 18:51:20 Voltage Offset Correction: v_off_ini = 2249.460 uV, v_off_fin = 2234.320 uV 18:51:20 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 70.845 uVs 18:51:25 I1X Average and RMS of Measurements VT = -96.829 +- 12.969 uVs I1X = 19.366 +- 2.594 uTm 18:51:27 Wire Position: X0 = 1.0000 mm, Y0 = 0.0000 mm 18:51:31 Integrated Voltage Measurement... 18:51:40 Voltage Offset Correction: v_off_ini = 2211.119 uV, v_off_fin = 2210.571 uV 18:51:40 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -194.221 uVs 18:51:40 Integrated Voltage Measurement... 18:51:49 Voltage Offset Correction: v_off_ini = 2231.604 uV, v_off_fin = 2232.836 uV 18:51:49 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 214.280 uVs 18:51:49 Integrated Voltage Measurement... 18:51:58 Voltage Offset Correction: v_off_ini = 2197.262 uV, v_off_fin = 2191.454 uV 18:51:58 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -211.299 uVs 18:51:58 Integrated Voltage Measurement... 18:52:07 Voltage Offset Correction: v_off_ini = 2196.306 uV, v_off_fin = 2217.826 uV 18:52:07 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 200.076 uVs 18:52:07 Integrated Voltage Measurement... 18:52:16 Voltage Offset Correction: v_off_ini = 2217.793 uV, v_off_fin = 2217.047 uV 18:52:16 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -199.407 uVs 18:52:16 Integrated Voltage Measurement... 18:52:25 Voltage Offset Correction: v_off_ini = 2228.790 uV, v_off_fin = 2253.670 uV 18:52:25 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 192.511 uVs 18:52:29 I1Y Average and RMS of Measurements VT = -201.966 +- 4.288 uVs I1Y = -40.393 +- 0.858 uTm 18:52:34 Wire Position: X0 = 1.0000 mm, Y0 = 0.0000 mm 18:52:37 Integrated Voltage Measurement... 18:52:46 Voltage Offset Correction: v_off_ini = 2254.441 uV, v_off_fin = 2245.143 uV 18:52:46 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -67.252 uVs 18:52:46 Integrated Voltage Measurement... 18:52:55 Voltage Offset Correction: v_off_ini = 2239.454 uV, v_off_fin = 2271.281 uV 18:52:55 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 59.719 uVs 18:52:55 Integrated Voltage Measurement... 18:53:04 Voltage Offset Correction: v_off_ini = 2271.761 uV, v_off_fin = 2224.523 uV 18:53:04 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -74.336 uVs 18:53:04 Integrated Voltage Measurement... 18:53:13 Voltage Offset Correction: v_off_ini = 2258.875 uV, v_off_fin = 2257.605 uV 18:53:13 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 62.220 uVs 18:53:13 Integrated Voltage Measurement... 18:53:22 Voltage Offset Correction: v_off_ini = 2261.555 uV, v_off_fin = 2264.721 uV 18:53:22 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -100.513 uVs 18:53:22 Integrated Voltage Measurement... 18:53:31 Voltage Offset Correction: v_off_ini = 2266.000 uV, v_off_fin = 2269.504 uV 18:53:31 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 70.975 uVs 18:53:33 I2X Average and RMS of Measurements VT = -72.503 +- 9.565 uVs I2X = 6.670 +- 0.880 uTm^2 18:53:38 Wire Position: X0 = 1.0000 mm, Y0 = 0.0000 mm 18:53:41 Integrated Voltage Measurement... 18:53:50 Voltage Offset Correction: v_off_ini = 2250.585 uV, v_off_fin = 2233.303 uV 18:53:50 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -200.194 uVs 18:53:50 Integrated Voltage Measurement... 18:53:59 Voltage Offset Correction: v_off_ini = 2246.859 uV, v_off_fin = 2243.113 uV 18:53:59 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 186.523 uVs 18:53:59 Integrated Voltage Measurement... 18:54:08 Voltage Offset Correction: v_off_ini = 2262.121 uV, v_off_fin = 2261.021 uV 18:54:08 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -223.724 uVs 18:54:08 Integrated Voltage Measurement... 18:54:17 Voltage Offset Correction: v_off_ini = 2256.954 uV, v_off_fin = 2242.560 uV 18:54:17 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = 206.992 uVs 18:54:17 Integrated Voltage Measurement... 18:54:26 Voltage Offset Correction: v_off_ini = 2252.982 uV, v_off_fin = 2251.118 uV 18:54:26 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -198.388 uVs 18:54:26 Integrated Voltage Measurement... 18:54:35 Voltage Offset Correction: v_off_ini = 2251.374 uV, v_off_fin = 2248.520 uV 18:54:35 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 206.744 uVs 18:54:37 I2Y Average and RMS of Measurements VT = -203.761 +- 9.021 uVs I2Y = -18.746 +- 0.830 uTm^2 18:54:42 Wire Position: X0 = 2.0000 mm, Y0 = 0.0000 mm 18:54:46 Integrated Voltage Measurement... 18:54:55 Voltage Offset Correction: v_off_ini = 2302.041 uV, v_off_fin = 2282.974 uV 18:54:55 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.4998 mm dXY = -0.4999 mm VT = -63.282 uVs 18:54:55 Integrated Voltage Measurement... 18:55:04 Voltage Offset Correction: v_off_ini = 2270.625 uV, v_off_fin = 2258.070 uV 18:55:04 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 68.799 uVs 18:55:04 Integrated Voltage Measurement... 18:55:13 Voltage Offset Correction: v_off_ini = 2237.979 uV, v_off_fin = 2236.455 uV 18:55:13 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -56.489 uVs 18:55:13 Integrated Voltage Measurement... 18:55:22 Voltage Offset Correction: v_off_ini = 2219.463 uV, v_off_fin = 2215.490 uV 18:55:22 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 85.505 uVs 18:55:22 Integrated Voltage Measurement... 18:55:31 Voltage Offset Correction: v_off_ini = 2232.147 uV, v_off_fin = 2240.908 uV 18:55:31 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -70.613 uVs 18:55:31 Integrated Voltage Measurement... 18:55:40 Voltage Offset Correction: v_off_ini = 2237.649 uV, v_off_fin = 2243.881 uV 18:55:40 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 67.559 uVs 18:55:44 I1X Average and RMS of Measurements VT = -68.708 +- 2.041 uVs I1X = 13.742 +- 0.407 uTm 18:55:47 Wire Position: X0 = 2.0000 mm, Y0 = 0.0000 mm 18:55:51 Integrated Voltage Measurement... 18:56:00 Voltage Offset Correction: v_off_ini = 2242.505 uV, v_off_fin = 2233.484 uV 18:56:00 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -191.626 uVs 18:56:00 Integrated Voltage Measurement... 18:56:09 Voltage Offset Correction: v_off_ini = 2264.572 uV, v_off_fin = 2266.273 uV 18:56:09 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 201.513 uVs 18:56:09 Integrated Voltage Measurement... 18:56:18 Voltage Offset Correction: v_off_ini = 2251.294 uV, v_off_fin = 2245.518 uV 18:56:18 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -208.991 uVs 18:56:18 Integrated Voltage Measurement... 18:56:27 Voltage Offset Correction: v_off_ini = 2250.171 uV, v_off_fin = 2255.032 uV 18:56:27 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 197.547 uVs 18:56:27 Integrated Voltage Measurement... 18:56:36 Voltage Offset Correction: v_off_ini = 2238.156 uV, v_off_fin = 2238.488 uV 18:56:36 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -221.740 uVs 18:56:36 Integrated Voltage Measurement... 18:56:45 Voltage Offset Correction: v_off_ini = 2244.930 uV, v_off_fin = 2252.691 uV 18:56:45 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 225.589 uVs 18:56:49 I1Y Average and RMS of Measurements VT = -207.835 +- 11.523 uVs I1Y = -41.567 +- 2.305 uTm 18:56:54 Wire Position: X0 = 2.0000 mm, Y0 = 0.0000 mm 18:56:57 Integrated Voltage Measurement... 18:57:06 Voltage Offset Correction: v_off_ini = 2228.322 uV, v_off_fin = 2238.942 uV 18:57:06 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -60.749 uVs 18:57:06 Integrated Voltage Measurement... 18:57:15 Voltage Offset Correction: v_off_ini = 2247.285 uV, v_off_fin = 2264.611 uV 18:57:15 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 67.710 uVs 18:57:15 Integrated Voltage Measurement... 18:57:24 Voltage Offset Correction: v_off_ini = 2238.178 uV, v_off_fin = 2259.068 uV 18:57:24 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -17.308 uVs 18:57:24 Integrated Voltage Measurement... 18:57:33 Voltage Offset Correction: v_off_ini = 2252.405 uV, v_off_fin = 2252.648 uV 18:57:33 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 31.036 uVs 18:57:33 Integrated Voltage Measurement... 18:57:42 Voltage Offset Correction: v_off_ini = 2243.512 uV, v_off_fin = 2243.561 uV 18:57:42 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -36.528 uVs 18:57:42 Integrated Voltage Measurement... 18:57:51 Voltage Offset Correction: v_off_ini = 2240.458 uV, v_off_fin = 2252.122 uV 18:57:51 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 61.556 uVs 18:57:53 I2X Average and RMS of Measurements VT = -45.815 +- 16.512 uVs I2X = 4.215 +- 1.519 uTm^2 18:57:58 Wire Position: X0 = 2.0000 mm, Y0 = 0.0000 mm 18:58:01 Integrated Voltage Measurement... 18:58:10 Voltage Offset Correction: v_off_ini = 2235.438 uV, v_off_fin = 2241.071 uV 18:58:10 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -193.660 uVs 18:58:10 Integrated Voltage Measurement... 18:58:19 Voltage Offset Correction: v_off_ini = 2250.423 uV, v_off_fin = 2250.447 uV 18:58:19 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 198.685 uVs 18:58:19 Integrated Voltage Measurement... 18:58:28 Voltage Offset Correction: v_off_ini = 2246.094 uV, v_off_fin = 2244.362 uV 18:58:28 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -184.054 uVs 18:58:28 Integrated Voltage Measurement... 18:58:37 Voltage Offset Correction: v_off_ini = 2244.660 uV, v_off_fin = 2255.206 uV 18:58:37 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 201.503 uVs 18:58:37 Integrated Voltage Measurement... 18:58:46 Voltage Offset Correction: v_off_ini = 2233.902 uV, v_off_fin = 2246.040 uV 18:58:46 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -198.908 uVs 18:58:46 Integrated Voltage Measurement... 18:58:55 Voltage Offset Correction: v_off_ini = 2219.281 uV, v_off_fin = 2221.575 uV 18:58:55 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 183.004 uVs 18:58:57 I2Y Average and RMS of Measurements VT = -193.302 +- 2.162 uVs I2Y = -17.784 +- 0.199 uTm^2 18:59:05 Wire Position: X0 = 0.0000 mm, Y0 = -1.0000 mm 18:59:09 Integrated Voltage Measurement... 18:59:18 Voltage Offset Correction: v_off_ini = 2198.314 uV, v_off_fin = 2224.103 uV 18:59:18 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -86.699 uVs 18:59:18 Integrated Voltage Measurement... 18:59:27 Voltage Offset Correction: v_off_ini = 2227.378 uV, v_off_fin = 2221.107 uV 18:59:27 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 125.828 uVs 18:59:27 Integrated Voltage Measurement... 18:59:36 Voltage Offset Correction: v_off_ini = 2214.980 uV, v_off_fin = 2211.172 uV 18:59:36 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -94.574 uVs 18:59:36 Integrated Voltage Measurement... 18:59:45 Voltage Offset Correction: v_off_ini = 2215.073 uV, v_off_fin = 2223.209 uV 18:59:45 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 139.448 uVs 18:59:45 Integrated Voltage Measurement... 18:59:54 Voltage Offset Correction: v_off_ini = 2234.586 uV, v_off_fin = 2234.086 uV 18:59:54 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -112.783 uVs 18:59:54 Integrated Voltage Measurement... 19:00:03 Voltage Offset Correction: v_off_ini = 2247.531 uV, v_off_fin = 2247.169 uV 19:00:03 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 110.787 uVs 19:00:07 I1X Average and RMS of Measurements VT = -111.687 +- 4.388 uVs I1X = 22.338 +- 0.877 uTm 19:00:10 Wire Position: X0 = 0.0000 mm, Y0 = -1.0000 mm 19:00:14 Integrated Voltage Measurement... 19:00:23 Voltage Offset Correction: v_off_ini = 2257.092 uV, v_off_fin = 2238.720 uV 19:00:23 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -186.727 uVs 19:00:23 Integrated Voltage Measurement... 19:00:32 Voltage Offset Correction: v_off_ini = 2266.212 uV, v_off_fin = 2281.164 uV 19:00:32 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 229.480 uVs 19:00:32 Integrated Voltage Measurement... 19:00:41 Voltage Offset Correction: v_off_ini = 2271.358 uV, v_off_fin = 2247.783 uV 19:00:41 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -192.860 uVs 19:00:41 Integrated Voltage Measurement... 19:00:50 Voltage Offset Correction: v_off_ini = 2267.742 uV, v_off_fin = 2265.502 uV 19:00:50 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 218.375 uVs 19:00:50 Integrated Voltage Measurement... 19:00:59 Voltage Offset Correction: v_off_ini = 2272.767 uV, v_off_fin = 2275.702 uV 19:00:59 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -186.228 uVs 19:00:59 Integrated Voltage Measurement... 19:01:08 Voltage Offset Correction: v_off_ini = 2273.162 uV, v_off_fin = 2285.628 uV 19:01:08 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 230.503 uVs 19:01:12 I1Y Average and RMS of Measurements VT = -207.362 +- 1.238 uVs I1Y = -41.472 +- 0.248 uTm 19:01:15 Wire Position: X0 = 0.0000 mm, Y0 = -1.0000 mm 19:01:17 Integrated Voltage Measurement... 19:01:26 Voltage Offset Correction: v_off_ini = 2271.024 uV, v_off_fin = 2276.662 uV 19:01:26 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -109.170 uVs 19:01:26 Integrated Voltage Measurement... 19:01:35 Voltage Offset Correction: v_off_ini = 2283.013 uV, v_off_fin = 2267.635 uV 19:01:35 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 104.720 uVs 19:01:35 Integrated Voltage Measurement... 19:01:44 Voltage Offset Correction: v_off_ini = 2265.484 uV, v_off_fin = 2256.562 uV 19:01:44 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -93.894 uVs 19:01:44 Integrated Voltage Measurement... 19:01:53 Voltage Offset Correction: v_off_ini = 2267.159 uV, v_off_fin = 2294.354 uV 19:01:53 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 120.626 uVs 19:01:53 Integrated Voltage Measurement... 19:02:02 Voltage Offset Correction: v_off_ini = 2283.874 uV, v_off_fin = 2283.663 uV 19:02:02 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -113.517 uVs 19:02:02 Integrated Voltage Measurement... 19:02:11 Voltage Offset Correction: v_off_ini = 2287.168 uV, v_off_fin = 2301.630 uV 19:02:11 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 119.222 uVs 19:02:14 I2X Average and RMS of Measurements VT = -110.191 +- 4.370 uVs I2X = 10.138 +- 0.402 uTm^2 19:02:16 Wire Position: X0 = 0.0000 mm, Y0 = -1.0000 mm 19:02:19 Integrated Voltage Measurement... 19:02:28 Voltage Offset Correction: v_off_ini = 2283.112 uV, v_off_fin = 2267.657 uV 19:02:28 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -214.917 uVs 19:02:28 Integrated Voltage Measurement... 19:02:37 Voltage Offset Correction: v_off_ini = 2275.919 uV, v_off_fin = 2261.897 uV 19:02:37 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 234.157 uVs 19:02:37 Integrated Voltage Measurement... 19:02:46 Voltage Offset Correction: v_off_ini = 2257.156 uV, v_off_fin = 2262.362 uV 19:02:46 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -215.627 uVs 19:02:46 Integrated Voltage Measurement... 19:02:55 Voltage Offset Correction: v_off_ini = 2278.390 uV, v_off_fin = 2291.547 uV 19:02:55 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 216.110 uVs 19:02:55 Integrated Voltage Measurement... 19:03:04 Voltage Offset Correction: v_off_ini = 2263.794 uV, v_off_fin = 2251.258 uV 19:03:04 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -223.647 uVs 19:03:04 Integrated Voltage Measurement... 19:03:13 Voltage Offset Correction: v_off_ini = 2281.643 uV, v_off_fin = 2267.876 uV 19:03:13 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 202.698 uVs 19:03:15 I2Y Average and RMS of Measurements VT = -217.859 +- 4.849 uVs I2Y = -20.043 +- 0.446 uTm^2 19:03:19 Wire Position: X0 = 0.0000 mm, Y0 = -0.5000 mm 19:03:23 Integrated Voltage Measurement... 19:03:32 Voltage Offset Correction: v_off_ini = 2241.372 uV, v_off_fin = 2222.180 uV 19:03:32 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -116.005 uVs 19:03:32 Integrated Voltage Measurement... 19:03:41 Voltage Offset Correction: v_off_ini = 2228.966 uV, v_off_fin = 2239.229 uV 19:03:41 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 120.602 uVs 19:03:41 Integrated Voltage Measurement... 19:03:50 Voltage Offset Correction: v_off_ini = 2238.632 uV, v_off_fin = 2216.362 uV 19:03:50 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -121.180 uVs 19:03:50 Integrated Voltage Measurement... 19:03:59 Voltage Offset Correction: v_off_ini = 2229.887 uV, v_off_fin = 2241.060 uV 19:03:59 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 111.962 uVs 19:03:59 Integrated Voltage Measurement... 19:04:08 Voltage Offset Correction: v_off_ini = 2224.233 uV, v_off_fin = 2223.127 uV 19:04:08 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -111.464 uVs 19:04:08 Integrated Voltage Measurement... 19:04:17 Voltage Offset Correction: v_off_ini = 2213.003 uV, v_off_fin = 2229.078 uV 19:04:17 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 107.239 uVs 19:04:22 I1X Average and RMS of Measurements VT = -114.742 +- 3.877 uVs I1X = 22.949 +- 0.775 uTm 19:04:24 Wire Position: X0 = 0.0000 mm, Y0 = -0.5000 mm 19:04:28 Integrated Voltage Measurement... 19:04:37 Voltage Offset Correction: v_off_ini = 2236.768 uV, v_off_fin = 2231.255 uV 19:04:37 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -198.779 uVs 19:04:37 Integrated Voltage Measurement... 19:04:46 Voltage Offset Correction: v_off_ini = 2230.261 uV, v_off_fin = 2236.096 uV 19:04:46 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 199.292 uVs 19:04:46 Integrated Voltage Measurement... 19:04:55 Voltage Offset Correction: v_off_ini = 2203.273 uV, v_off_fin = 2182.517 uV 19:04:55 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -200.195 uVs 19:04:55 Integrated Voltage Measurement... 19:05:04 Voltage Offset Correction: v_off_ini = 2184.197 uV, v_off_fin = 2185.213 uV 19:05:04 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 176.620 uVs 19:05:04 Integrated Voltage Measurement... 19:05:13 Voltage Offset Correction: v_off_ini = 2182.678 uV, v_off_fin = 2179.880 uV 19:05:13 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -180.415 uVs 19:05:13 Integrated Voltage Measurement... 19:05:22 Voltage Offset Correction: v_off_ini = 2195.256 uV, v_off_fin = 2201.759 uV 19:05:22 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 188.283 uVs 19:05:26 I1Y Average and RMS of Measurements VT = -190.597 +- 6.192 uVs I1Y = -38.119 +- 1.238 uTm 19:05:29 Wire Position: X0 = 0.0000 mm, Y0 = -0.5000 mm 19:05:31 Integrated Voltage Measurement... 19:05:40 Voltage Offset Correction: v_off_ini = 2202.280 uV, v_off_fin = 2191.873 uV 19:05:40 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -124.139 uVs 19:05:40 Integrated Voltage Measurement... 19:05:49 Voltage Offset Correction: v_off_ini = 2197.949 uV, v_off_fin = 2204.173 uV 19:05:49 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 83.101 uVs 19:05:49 Integrated Voltage Measurement... 19:05:58 Voltage Offset Correction: v_off_ini = 2188.800 uV, v_off_fin = 2177.018 uV 19:05:58 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -95.171 uVs 19:05:58 Integrated Voltage Measurement... 19:06:07 Voltage Offset Correction: v_off_ini = 2184.123 uV, v_off_fin = 2169.368 uV 19:06:07 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 78.711 uVs 19:06:07 Integrated Voltage Measurement... 19:06:16 Voltage Offset Correction: v_off_ini = 2167.499 uV, v_off_fin = 2169.367 uV 19:06:16 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -109.749 uVs 19:06:16 Integrated Voltage Measurement... 19:06:25 Voltage Offset Correction: v_off_ini = 2186.469 uV, v_off_fin = 2207.181 uV 19:06:25 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 85.710 uVs 19:06:28 I2X Average and RMS of Measurements VT = -96.097 +- 6.906 uVs I2X = 8.841 +- 0.636 uTm^2 19:06:30 Wire Position: X0 = 0.0000 mm, Y0 = -0.5000 mm 19:06:33 Integrated Voltage Measurement... 19:06:42 Voltage Offset Correction: v_off_ini = 2219.364 uV, v_off_fin = 2212.183 uV 19:06:42 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -203.767 uVs 19:06:42 Integrated Voltage Measurement... 19:06:51 Voltage Offset Correction: v_off_ini = 2223.860 uV, v_off_fin = 2218.500 uV 19:06:51 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 208.448 uVs 19:06:51 Integrated Voltage Measurement... 19:07:00 Voltage Offset Correction: v_off_ini = 2195.939 uV, v_off_fin = 2196.053 uV 19:07:00 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -207.054 uVs 19:07:00 Integrated Voltage Measurement... 19:07:09 Voltage Offset Correction: v_off_ini = 2225.180 uV, v_off_fin = 2223.457 uV 19:07:09 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 197.786 uVs 19:07:09 Integrated Voltage Measurement... 19:07:18 Voltage Offset Correction: v_off_ini = 2199.488 uV, v_off_fin = 2207.079 uV 19:07:18 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -185.430 uVs 19:07:18 Integrated Voltage Measurement... 19:07:27 Voltage Offset Correction: v_off_ini = 2198.661 uV, v_off_fin = 2216.452 uV 19:07:27 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 220.919 uVs 19:07:29 I2Y Average and RMS of Measurements VT = -203.901 +- 1.590 uVs I2Y = -18.759 +- 0.146 uTm^2 19:07:33 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 19:07:38 Integrated Voltage Measurement... 19:07:47 Voltage Offset Correction: v_off_ini = 2220.972 uV, v_off_fin = 2224.156 uV 19:07:47 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -112.571 uVs 19:07:47 Integrated Voltage Measurement... 19:07:56 Voltage Offset Correction: v_off_ini = 2222.187 uV, v_off_fin = 2213.573 uV 19:07:56 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 113.919 uVs 19:07:56 Integrated Voltage Measurement... 19:08:05 Voltage Offset Correction: v_off_ini = 2207.555 uV, v_off_fin = 2207.395 uV 19:08:05 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -145.494 uVs 19:08:05 Integrated Voltage Measurement... 19:08:14 Voltage Offset Correction: v_off_ini = 2220.373 uV, v_off_fin = 2228.612 uV 19:08:14 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = 101.811 uVs 19:08:14 Integrated Voltage Measurement... 19:08:23 Voltage Offset Correction: v_off_ini = 2198.099 uV, v_off_fin = 2175.938 uV 19:08:23 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -108.380 uVs 19:08:23 Integrated Voltage Measurement... 19:08:32 Voltage Offset Correction: v_off_ini = 2182.359 uV, v_off_fin = 2190.207 uV 19:08:32 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 118.731 uVs 19:08:36 I1X Average and RMS of Measurements VT = -116.818 +- 4.835 uVs I1X = 23.364 +- 0.967 uTm 19:08:38 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 19:08:43 Integrated Voltage Measurement... 19:08:51 Voltage Offset Correction: v_off_ini = 2221.371 uV, v_off_fin = 2195.373 uV 19:08:51 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -195.020 uVs 19:08:51 Integrated Voltage Measurement... 19:09:00 Voltage Offset Correction: v_off_ini = 2216.704 uV, v_off_fin = 2231.014 uV 19:09:00 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 194.450 uVs 19:09:00 Integrated Voltage Measurement... 19:09:09 Voltage Offset Correction: v_off_ini = 2220.610 uV, v_off_fin = 2201.422 uV 19:09:09 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = -198.009 uVs 19:09:09 Integrated Voltage Measurement... 19:09:18 Voltage Offset Correction: v_off_ini = 2193.116 uV, v_off_fin = 2207.962 uV 19:09:18 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 202.736 uVs 19:09:18 Integrated Voltage Measurement... 19:09:27 Voltage Offset Correction: v_off_ini = 2210.890 uV, v_off_fin = 2199.823 uV 19:09:27 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -194.776 uVs 19:09:27 Integrated Voltage Measurement... 19:09:36 Voltage Offset Correction: v_off_ini = 2208.575 uV, v_off_fin = 2204.308 uV 19:09:36 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 187.542 uVs 19:09:40 I1Y Average and RMS of Measurements VT = -195.422 +- 3.793 uVs I1Y = -39.085 +- 0.759 uTm 19:09:43 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 19:09:46 Integrated Voltage Measurement... 19:09:55 Voltage Offset Correction: v_off_ini = 2221.243 uV, v_off_fin = 2217.987 uV 19:09:55 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -101.920 uVs 19:09:55 Integrated Voltage Measurement... 19:10:04 Voltage Offset Correction: v_off_ini = 2235.981 uV, v_off_fin = 2233.732 uV 19:10:04 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 103.080 uVs 19:10:04 Integrated Voltage Measurement... 19:10:13 Voltage Offset Correction: v_off_ini = 2211.581 uV, v_off_fin = 2200.055 uV 19:10:13 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -100.605 uVs 19:10:13 Integrated Voltage Measurement... 19:10:22 Voltage Offset Correction: v_off_ini = 2211.273 uV, v_off_fin = 2219.053 uV 19:10:22 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 98.689 uVs 19:10:22 Integrated Voltage Measurement... 19:10:31 Voltage Offset Correction: v_off_ini = 2218.127 uV, v_off_fin = 2207.645 uV 19:10:31 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -106.408 uVs 19:10:31 Integrated Voltage Measurement... 19:10:40 Voltage Offset Correction: v_off_ini = 2197.440 uV, v_off_fin = 2204.727 uV 19:10:40 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 101.582 uVs 19:10:42 I2X Average and RMS of Measurements VT = -102.047 +- 1.803 uVs I2X = 9.388 +- 0.166 uTm^2 19:10:44 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 19:10:47 Integrated Voltage Measurement... 19:10:56 Voltage Offset Correction: v_off_ini = 2236.216 uV, v_off_fin = 2237.189 uV 19:10:56 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -205.180 uVs 19:10:56 Integrated Voltage Measurement... 19:11:05 Voltage Offset Correction: v_off_ini = 2222.787 uV, v_off_fin = 2204.896 uV 19:11:05 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 185.477 uVs 19:11:05 Integrated Voltage Measurement... 19:11:14 Voltage Offset Correction: v_off_ini = 2200.882 uV, v_off_fin = 2211.302 uV 19:11:14 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -192.325 uVs 19:11:14 Integrated Voltage Measurement... 19:11:23 Voltage Offset Correction: v_off_ini = 2222.206 uV, v_off_fin = 2214.848 uV 19:11:23 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 180.971 uVs 19:11:23 Integrated Voltage Measurement... 19:11:32 Voltage Offset Correction: v_off_ini = 2220.339 uV, v_off_fin = 2212.508 uV 19:11:32 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -192.173 uVs 19:11:32 Integrated Voltage Measurement... 19:11:41 Voltage Offset Correction: v_off_ini = 2211.624 uV, v_off_fin = 2213.223 uV 19:11:41 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 172.288 uVs 19:11:44 I2Y Average and RMS of Measurements VT = -188.069 +- 5.441 uVs I2Y = -17.302 +- 0.501 uTm^2 19:11:48 Wire Position: X0 = 0.0000 mm, Y0 = 0.5000 mm 19:11:52 Integrated Voltage Measurement... 19:12:01 Voltage Offset Correction: v_off_ini = 2201.139 uV, v_off_fin = 2190.965 uV 19:12:01 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -102.540 uVs 19:12:01 Integrated Voltage Measurement... 19:12:10 Voltage Offset Correction: v_off_ini = 2199.598 uV, v_off_fin = 2191.326 uV 19:12:10 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 102.458 uVs 19:12:10 Integrated Voltage Measurement... 19:12:19 Voltage Offset Correction: v_off_ini = 2185.818 uV, v_off_fin = 2172.946 uV 19:12:19 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -105.257 uVs 19:12:19 Integrated Voltage Measurement... 19:12:28 Voltage Offset Correction: v_off_ini = 2190.280 uV, v_off_fin = 2220.169 uV 19:12:28 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 96.422 uVs 19:12:28 Integrated Voltage Measurement... 19:12:37 Voltage Offset Correction: v_off_ini = 2213.034 uV, v_off_fin = 2208.360 uV 19:12:37 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -106.733 uVs 19:12:37 Integrated Voltage Measurement... 19:12:46 Voltage Offset Correction: v_off_ini = 2201.786 uV, v_off_fin = 2208.054 uV 19:12:46 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 128.879 uVs 19:12:50 I1X Average and RMS of Measurements VT = -107.048 +- 7.637 uVs I1X = 21.410 +- 1.527 uTm 19:12:52 Wire Position: X0 = 0.0000 mm, Y0 = 0.5000 mm 19:12:57 Integrated Voltage Measurement... 19:13:06 Voltage Offset Correction: v_off_ini = 2198.097 uV, v_off_fin = 2202.375 uV 19:13:06 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -159.691 uVs 19:13:06 Integrated Voltage Measurement... 19:13:15 Voltage Offset Correction: v_off_ini = 2200.810 uV, v_off_fin = 2221.629 uV 19:13:15 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 179.985 uVs 19:13:15 Integrated Voltage Measurement... 19:13:24 Voltage Offset Correction: v_off_ini = 2211.661 uV, v_off_fin = 2198.200 uV 19:13:24 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -166.716 uVs 19:13:24 Integrated Voltage Measurement... 19:13:33 Voltage Offset Correction: v_off_ini = 2225.476 uV, v_off_fin = 2222.408 uV 19:13:33 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 173.629 uVs 19:13:33 Integrated Voltage Measurement... 19:13:42 Voltage Offset Correction: v_off_ini = 2204.895 uV, v_off_fin = 2208.175 uV 19:13:42 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -161.790 uVs 19:13:42 Integrated Voltage Measurement... 19:13:51 Voltage Offset Correction: v_off_ini = 2228.795 uV, v_off_fin = 2210.878 uV 19:13:51 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 175.175 uVs 19:13:55 I1Y Average and RMS of Measurements VT = -169.498 +- 0.731 uVs I1Y = -33.898 +- 0.146 uTm 19:13:57 Wire Position: X0 = 0.0000 mm, Y0 = 0.5000 mm 19:14:00 Integrated Voltage Measurement... 19:14:09 Voltage Offset Correction: v_off_ini = 2199.354 uV, v_off_fin = 2199.365 uV 19:14:09 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -72.547 uVs 19:14:09 Integrated Voltage Measurement... 19:14:18 Voltage Offset Correction: v_off_ini = 2204.249 uV, v_off_fin = 2208.826 uV 19:14:18 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 80.576 uVs 19:14:18 Integrated Voltage Measurement... 19:14:27 Voltage Offset Correction: v_off_ini = 2197.977 uV, v_off_fin = 2184.234 uV 19:14:27 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -104.237 uVs 19:14:27 Integrated Voltage Measurement... 19:14:36 Voltage Offset Correction: v_off_ini = 2194.934 uV, v_off_fin = 2197.229 uV 19:14:36 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 108.256 uVs 19:14:36 Integrated Voltage Measurement... 19:14:45 Voltage Offset Correction: v_off_ini = 2206.006 uV, v_off_fin = 2202.434 uV 19:14:45 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -63.588 uVs 19:14:45 Integrated Voltage Measurement... 19:14:54 Voltage Offset Correction: v_off_ini = 2191.128 uV, v_off_fin = 2203.974 uV 19:14:54 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 106.990 uVs 19:14:56 I2X Average and RMS of Measurements VT = -89.366 +- 12.457 uVs I2X = 8.222 +- 1.146 uTm^2 19:14:59 Wire Position: X0 = 0.0000 mm, Y0 = 0.5000 mm 19:15:01 Integrated Voltage Measurement... 19:15:10 Voltage Offset Correction: v_off_ini = 2182.604 uV, v_off_fin = 2183.228 uV 19:15:10 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -155.848 uVs 19:15:10 Integrated Voltage Measurement... 19:15:19 Voltage Offset Correction: v_off_ini = 2186.174 uV, v_off_fin = 2210.971 uV 19:15:19 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 165.780 uVs 19:15:19 Integrated Voltage Measurement... 19:15:28 Voltage Offset Correction: v_off_ini = 2182.665 uV, v_off_fin = 2154.246 uV 19:15:28 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -189.385 uVs 19:15:28 Integrated Voltage Measurement... 19:15:37 Voltage Offset Correction: v_off_ini = 2171.897 uV, v_off_fin = 2196.601 uV 19:15:37 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 170.895 uVs 19:15:37 Integrated Voltage Measurement... 19:15:46 Voltage Offset Correction: v_off_ini = 2181.853 uV, v_off_fin = 2187.397 uV 19:15:46 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -180.851 uVs 19:15:46 Integrated Voltage Measurement... 19:15:55 Voltage Offset Correction: v_off_ini = 2184.360 uV, v_off_fin = 2177.859 uV 19:15:55 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 188.524 uVs 19:15:58 I2Y Average and RMS of Measurements VT = -175.214 +- 10.350 uVs I2Y = -16.120 +- 0.952 uTm^2 19:16:02 Wire Position: X0 = 0.0000 mm, Y0 = 1.0000 mm 19:16:06 Integrated Voltage Measurement... 19:16:15 Voltage Offset Correction: v_off_ini = 2189.105 uV, v_off_fin = 2192.149 uV 19:16:15 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -86.060 uVs 19:16:15 Integrated Voltage Measurement... 19:16:24 Voltage Offset Correction: v_off_ini = 2184.586 uV, v_off_fin = 2194.824 uV 19:16:24 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 103.795 uVs 19:16:24 Integrated Voltage Measurement... 19:16:33 Voltage Offset Correction: v_off_ini = 2193.853 uV, v_off_fin = 2176.327 uV 19:16:33 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -73.484 uVs 19:16:33 Integrated Voltage Measurement... 19:16:42 Voltage Offset Correction: v_off_ini = 2167.307 uV, v_off_fin = 2163.407 uV 19:16:42 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 121.693 uVs 19:16:42 Integrated Voltage Measurement... 19:16:51 Voltage Offset Correction: v_off_ini = 2167.157 uV, v_off_fin = 2179.166 uV 19:16:51 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -102.951 uVs 19:16:51 Integrated Voltage Measurement... 19:17:00 Voltage Offset Correction: v_off_ini = 2178.348 uV, v_off_fin = 2179.974 uV 19:17:00 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 82.211 uVs 19:17:04 I1X Average and RMS of Measurements VT = -95.032 +- 2.046 uVs I1X = 19.007 +- 0.409 uTm 19:17:07 Wire Position: X0 = 0.0000 mm, Y0 = 1.0000 mm 19:17:11 Integrated Voltage Measurement... 19:17:20 Voltage Offset Correction: v_off_ini = 2190.733 uV, v_off_fin = 2174.952 uV 19:17:20 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -167.778 uVs 19:17:20 Integrated Voltage Measurement... 19:17:29 Voltage Offset Correction: v_off_ini = 2189.110 uV, v_off_fin = 2184.004 uV 19:17:29 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 168.350 uVs 19:17:29 Integrated Voltage Measurement... 19:17:38 Voltage Offset Correction: v_off_ini = 2171.375 uV, v_off_fin = 2172.338 uV 19:17:38 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -175.035 uVs 19:17:38 Integrated Voltage Measurement... 19:17:47 Voltage Offset Correction: v_off_ini = 2192.747 uV, v_off_fin = 2202.886 uV 19:17:47 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 159.211 uVs 19:17:47 Integrated Voltage Measurement... 19:17:56 Voltage Offset Correction: v_off_ini = 2182.217 uV, v_off_fin = 2179.601 uV 19:17:56 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -147.241 uVs 19:17:56 Integrated Voltage Measurement... 19:18:05 Voltage Offset Correction: v_off_ini = 2164.848 uV, v_off_fin = 2163.431 uV 19:18:05 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 169.292 uVs 19:18:09 I1Y Average and RMS of Measurements VT = -164.485 +- 4.414 uVs I1Y = -32.896 +- 0.883 uTm 19:18:11 Wire Position: X0 = 0.0000 mm, Y0 = 1.0000 mm 19:18:14 Integrated Voltage Measurement... 19:18:23 Voltage Offset Correction: v_off_ini = 2183.665 uV, v_off_fin = 2180.218 uV 19:18:23 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -83.698 uVs 19:18:23 Integrated Voltage Measurement... 19:18:32 Voltage Offset Correction: v_off_ini = 2177.004 uV, v_off_fin = 2203.297 uV 19:18:32 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 84.839 uVs 19:18:32 Integrated Voltage Measurement... 19:18:41 Voltage Offset Correction: v_off_ini = 2202.246 uV, v_off_fin = 2169.042 uV 19:18:41 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -75.167 uVs 19:18:41 Integrated Voltage Measurement... 19:18:50 Voltage Offset Correction: v_off_ini = 2174.347 uV, v_off_fin = 2186.244 uV 19:18:50 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 86.422 uVs 19:18:50 Integrated Voltage Measurement... 19:18:59 Voltage Offset Correction: v_off_ini = 2182.150 uV, v_off_fin = 2176.212 uV 19:18:59 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -90.424 uVs 19:18:59 Integrated Voltage Measurement... 19:19:08 Voltage Offset Correction: v_off_ini = 2167.760 uV, v_off_fin = 2177.948 uV 19:19:08 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 73.448 uVs 19:19:10 I2X Average and RMS of Measurements VT = -82.333 +- 1.446 uVs I2X = 7.575 +- 0.133 uTm^2 19:19:13 Wire Position: X0 = 0.0000 mm, Y0 = 1.0000 mm 19:19:16 Integrated Voltage Measurement... 19:19:25 Voltage Offset Correction: v_off_ini = 2181.230 uV, v_off_fin = 2173.699 uV 19:19:25 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -157.987 uVs 19:19:25 Integrated Voltage Measurement... 19:19:34 Voltage Offset Correction: v_off_ini = 2184.618 uV, v_off_fin = 2179.565 uV 19:19:34 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 170.509 uVs 19:19:34 Integrated Voltage Measurement... 19:19:43 Voltage Offset Correction: v_off_ini = 2186.468 uV, v_off_fin = 2173.476 uV 19:19:43 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -142.639 uVs 19:19:43 Integrated Voltage Measurement... 19:19:52 Voltage Offset Correction: v_off_ini = 2201.947 uV, v_off_fin = 2192.043 uV 19:19:52 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 168.910 uVs 19:19:52 Integrated Voltage Measurement... 19:20:01 Voltage Offset Correction: v_off_ini = 2181.943 uV, v_off_fin = 2186.284 uV 19:20:01 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -172.481 uVs 19:20:01 Integrated Voltage Measurement... 19:20:10 Voltage Offset Correction: v_off_ini = 2202.975 uV, v_off_fin = 2195.163 uV 19:20:10 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 153.280 uVs 19:20:12 I2Y Average and RMS of Measurements VT = -160.968 +- 3.714 uVs I2Y = -14.809 +- 0.342 uTm^2 19:20:29 Wire Position: X0 = -2.0000 mm, Y0 = 0.0000 mm 19:20:33 Integrated Voltage Measurement... 19:20:42 Voltage Offset Correction: v_off_ini = 2212.117 uV, v_off_fin = 2198.609 uV 19:20:42 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -147.854 uVs 19:20:42 Integrated Voltage Measurement... 19:20:51 Voltage Offset Correction: v_off_ini = 2203.691 uV, v_off_fin = 2198.856 uV 19:20:51 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 170.341 uVs 19:20:51 Integrated Voltage Measurement... 19:21:00 Voltage Offset Correction: v_off_ini = 2193.802 uV, v_off_fin = 2198.679 uV 19:21:00 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.4999 mm dXY = -0.4999 mm VT = -162.502 uVs 19:21:00 Integrated Voltage Measurement... 19:21:09 Voltage Offset Correction: v_off_ini = 2187.726 uV, v_off_fin = 2187.878 uV 19:21:09 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = 170.650 uVs 19:21:09 Integrated Voltage Measurement... 19:21:18 Voltage Offset Correction: v_off_ini = 2176.175 uV, v_off_fin = 2176.311 uV 19:21:18 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -150.542 uVs 19:21:18 Integrated Voltage Measurement... 19:21:27 Voltage Offset Correction: v_off_ini = 2184.595 uV, v_off_fin = 2200.968 uV 19:21:27 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 143.924 uVs 19:21:31 I1X Average and RMS of Measurements VT = -157.635 +- 7.964 uVs I1X = 31.529 +- 1.595 uTm 19:21:34 Wire Position: X0 = -2.0000 mm, Y0 = 0.0000 mm 19:21:38 Integrated Voltage Measurement... 19:21:47 Voltage Offset Correction: v_off_ini = 2190.807 uV, v_off_fin = 2184.045 uV 19:21:47 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -150.103 uVs 19:21:47 Integrated Voltage Measurement... 19:21:56 Voltage Offset Correction: v_off_ini = 2192.324 uV, v_off_fin = 2186.625 uV 19:21:56 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 139.217 uVs 19:21:56 Integrated Voltage Measurement... 19:22:05 Voltage Offset Correction: v_off_ini = 2198.565 uV, v_off_fin = 2188.453 uV 19:22:05 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -157.646 uVs 19:22:05 Integrated Voltage Measurement... 19:22:14 Voltage Offset Correction: v_off_ini = 2189.032 uV, v_off_fin = 2194.349 uV 19:22:14 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 140.599 uVs 19:22:14 Integrated Voltage Measurement... 19:22:23 Voltage Offset Correction: v_off_ini = 2175.365 uV, v_off_fin = 2167.437 uV 19:22:23 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -157.265 uVs 19:22:23 Integrated Voltage Measurement... 19:22:32 Voltage Offset Correction: v_off_ini = 2207.632 uV, v_off_fin = 2207.378 uV 19:22:32 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 153.310 uVs 19:22:36 I1Y Average and RMS of Measurements VT = -149.690 +- 4.357 uVs I1Y = -29.938 +- 0.871 uTm 19:22:41 Wire Position: X0 = -2.0000 mm, Y0 = 0.0000 mm 19:22:44 Integrated Voltage Measurement... 19:22:53 Voltage Offset Correction: v_off_ini = 2191.153 uV, v_off_fin = 2192.870 uV 19:22:53 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -147.241 uVs 19:22:53 Integrated Voltage Measurement... 19:23:02 Voltage Offset Correction: v_off_ini = 2217.465 uV, v_off_fin = 2224.208 uV 19:23:02 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 135.175 uVs 19:23:02 Integrated Voltage Measurement... 19:23:11 Voltage Offset Correction: v_off_ini = 2224.214 uV, v_off_fin = 2209.915 uV 19:23:11 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -138.379 uVs 19:23:11 Integrated Voltage Measurement... 19:23:20 Voltage Offset Correction: v_off_ini = 2205.675 uV, v_off_fin = 2252.310 uV 19:23:20 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 132.334 uVs 19:23:20 Integrated Voltage Measurement... 19:23:29 Voltage Offset Correction: v_off_ini = 2229.941 uV, v_off_fin = 2209.094 uV 19:23:29 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -152.715 uVs 19:23:29 Integrated Voltage Measurement... 19:23:38 Voltage Offset Correction: v_off_ini = 2207.383 uV, v_off_fin = 2207.551 uV 19:23:38 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 142.873 uVs 19:23:40 I2X Average and RMS of Measurements VT = -141.453 +- 5.081 uVs I2X = 13.014 +- 0.467 uTm^2 19:23:45 Wire Position: X0 = -2.0000 mm, Y0 = 0.0000 mm 19:23:48 Integrated Voltage Measurement... 19:23:57 Voltage Offset Correction: v_off_ini = 2180.400 uV, v_off_fin = 2165.731 uV 19:23:57 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -157.580 uVs 19:23:57 Integrated Voltage Measurement... 19:24:06 Voltage Offset Correction: v_off_ini = 2184.350 uV, v_off_fin = 2185.130 uV 19:24:06 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 142.229 uVs 19:24:06 Integrated Voltage Measurement... 19:24:15 Voltage Offset Correction: v_off_ini = 2178.679 uV, v_off_fin = 2177.123 uV 19:24:15 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -154.032 uVs 19:24:15 Integrated Voltage Measurement... 19:24:24 Voltage Offset Correction: v_off_ini = 2173.429 uV, v_off_fin = 2155.651 uV 19:24:24 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 182.202 uVs 19:24:24 Integrated Voltage Measurement... 19:24:33 Voltage Offset Correction: v_off_ini = 2178.506 uV, v_off_fin = 2173.595 uV 19:24:33 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -158.027 uVs 19:24:33 Integrated Voltage Measurement... 19:24:42 Voltage Offset Correction: v_off_ini = 2162.647 uV, v_off_fin = 2166.768 uV 19:24:42 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 165.613 uVs 19:24:44 I2Y Average and RMS of Measurements VT = -159.947 +- 7.552 uVs I2Y = -14.715 +- 0.695 uTm^2 19:24:49 Wire Position: X0 = -1.0000 mm, Y0 = 0.0000 mm 19:24:53 Integrated Voltage Measurement... 19:25:02 Voltage Offset Correction: v_off_ini = 2176.124 uV, v_off_fin = 2174.465 uV 19:25:02 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4998 mm dXY = -0.4999 mm VT = -147.806 uVs 19:25:02 Integrated Voltage Measurement... 19:25:11 Voltage Offset Correction: v_off_ini = 2180.446 uV, v_off_fin = 2178.526 uV 19:25:11 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 139.915 uVs 19:25:11 Integrated Voltage Measurement... 19:25:20 Voltage Offset Correction: v_off_ini = 2174.795 uV, v_off_fin = 2163.981 uV 19:25:20 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -132.965 uVs 19:25:20 Integrated Voltage Measurement... 19:25:29 Voltage Offset Correction: v_off_ini = 2170.216 uV, v_off_fin = 2168.196 uV 19:25:29 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 120.946 uVs 19:25:29 Integrated Voltage Measurement... 19:25:38 Voltage Offset Correction: v_off_ini = 2150.643 uV, v_off_fin = 2146.414 uV 19:25:38 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -117.563 uVs 19:25:38 Integrated Voltage Measurement... 19:25:47 Voltage Offset Correction: v_off_ini = 2148.838 uV, v_off_fin = 2169.137 uV 19:25:47 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 120.221 uVs 19:25:51 I1X Average and RMS of Measurements VT = -129.903 +- 10.404 uVs I1X = 25.982 +- 2.082 uTm 19:25:54 Wire Position: X0 = -1.0000 mm, Y0 = 0.0000 mm 19:25:58 Integrated Voltage Measurement... 19:26:07 Voltage Offset Correction: v_off_ini = 2163.197 uV, v_off_fin = 2147.560 uV 19:26:07 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -140.718 uVs 19:26:07 Integrated Voltage Measurement... 19:26:16 Voltage Offset Correction: v_off_ini = 2159.113 uV, v_off_fin = 2149.800 uV 19:26:16 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 178.975 uVs 19:26:16 Integrated Voltage Measurement... 19:26:25 Voltage Offset Correction: v_off_ini = 2150.194 uV, v_off_fin = 2122.237 uV 19:26:25 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -160.546 uVs 19:26:25 Integrated Voltage Measurement... 19:26:34 Voltage Offset Correction: v_off_ini = 2119.401 uV, v_off_fin = 2141.556 uV 19:26:34 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 171.221 uVs 19:26:34 Integrated Voltage Measurement... 19:26:43 Voltage Offset Correction: v_off_ini = 2154.574 uV, v_off_fin = 2132.072 uV 19:26:43 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -153.300 uVs 19:26:43 Integrated Voltage Measurement... 19:26:52 Voltage Offset Correction: v_off_ini = 2130.510 uV, v_off_fin = 2142.063 uV 19:26:52 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 142.236 uVs 19:26:56 I1Y Average and RMS of Measurements VT = -157.832 +- 7.532 uVs I1Y = -31.566 +- 1.506 uTm 19:27:01 Wire Position: X0 = -1.0000 mm, Y0 = 0.0000 mm 19:27:04 Integrated Voltage Measurement... 19:27:13 Voltage Offset Correction: v_off_ini = 2106.377 uV, v_off_fin = 2102.629 uV 19:27:13 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -126.829 uVs 19:27:13 Integrated Voltage Measurement... 19:27:22 Voltage Offset Correction: v_off_ini = 2120.123 uV, v_off_fin = 2132.151 uV 19:27:22 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 128.777 uVs 19:27:22 Integrated Voltage Measurement... 19:27:31 Voltage Offset Correction: v_off_ini = 2116.345 uV, v_off_fin = 2120.105 uV 19:27:31 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -138.355 uVs 19:27:31 Integrated Voltage Measurement... 19:27:40 Voltage Offset Correction: v_off_ini = 2113.877 uV, v_off_fin = 2097.884 uV 19:27:40 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 126.871 uVs 19:27:40 Integrated Voltage Measurement... 19:27:49 Voltage Offset Correction: v_off_ini = 2089.252 uV, v_off_fin = 2078.494 uV 19:27:49 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -116.824 uVs 19:27:49 Integrated Voltage Measurement... 19:27:58 Voltage Offset Correction: v_off_ini = 2084.742 uV, v_off_fin = 2102.033 uV 19:27:58 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 103.060 uVs 19:28:00 I2X Average and RMS of Measurements VT = -123.453 +- 9.753 uVs I2X = 11.358 +- 0.897 uTm^2 19:28:05 Wire Position: X0 = -1.0000 mm, Y0 = 0.0000 mm 19:28:08 Integrated Voltage Measurement... 19:28:17 Voltage Offset Correction: v_off_ini = 2105.012 uV, v_off_fin = 2101.853 uV 19:28:17 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -142.813 uVs 19:28:17 Integrated Voltage Measurement... 19:28:26 Voltage Offset Correction: v_off_ini = 2080.666 uV, v_off_fin = 2080.833 uV 19:28:26 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 169.766 uVs 19:28:26 Integrated Voltage Measurement... 19:28:35 Voltage Offset Correction: v_off_ini = 2077.204 uV, v_off_fin = 2074.647 uV 19:28:35 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -159.223 uVs 19:28:35 Integrated Voltage Measurement... 19:28:44 Voltage Offset Correction: v_off_ini = 2085.878 uV, v_off_fin = 2081.869 uV 19:28:44 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 151.340 uVs 19:28:44 Integrated Voltage Measurement... 19:28:53 Voltage Offset Correction: v_off_ini = 2089.124 uV, v_off_fin = 2069.675 uV 19:28:53 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -149.491 uVs 19:28:53 Integrated Voltage Measurement... 19:29:02 Voltage Offset Correction: v_off_ini = 2084.375 uV, v_off_fin = 2080.582 uV 19:29:02 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 143.498 uVs 19:29:04 I2Y Average and RMS of Measurements VT = -152.688 +- 4.399 uVs I2Y = -14.047 +- 0.405 uTm^2 19:29:09 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 19:29:13 Integrated Voltage Measurement... 19:29:22 Voltage Offset Correction: v_off_ini = 2094.531 uV, v_off_fin = 2076.500 uV 19:29:22 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -134.778 uVs 19:29:22 Integrated Voltage Measurement... 19:29:31 Voltage Offset Correction: v_off_ini = 2066.310 uV, v_off_fin = 2076.405 uV 19:29:31 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 129.135 uVs 19:29:31 Integrated Voltage Measurement... 19:29:40 Voltage Offset Correction: v_off_ini = 2067.168 uV, v_off_fin = 2057.355 uV 19:29:40 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -116.867 uVs 19:29:40 Integrated Voltage Measurement... 19:29:49 Voltage Offset Correction: v_off_ini = 2092.366 uV, v_off_fin = 2094.331 uV 19:29:49 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 113.548 uVs 19:29:49 Integrated Voltage Measurement... 19:29:58 Voltage Offset Correction: v_off_ini = 2096.736 uV, v_off_fin = 2083.506 uV 19:29:58 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -110.280 uVs 19:29:58 Integrated Voltage Measurement... 19:30:07 Voltage Offset Correction: v_off_ini = 2099.360 uV, v_off_fin = 2111.921 uV 19:30:07 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 110.962 uVs 19:30:11 I1X Average and RMS of Measurements VT = -119.262 +- 9.170 uVs I1X = 23.853 +- 1.834 uTm 19:30:14 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 19:30:18 Integrated Voltage Measurement... 19:30:27 Voltage Offset Correction: v_off_ini = 2090.507 uV, v_off_fin = 2066.740 uV 19:30:27 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -177.320 uVs 19:30:27 Integrated Voltage Measurement... 19:30:36 Voltage Offset Correction: v_off_ini = 2114.733 uV, v_off_fin = 2130.813 uV 19:30:36 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 171.148 uVs 19:30:36 Integrated Voltage Measurement... 19:30:45 Voltage Offset Correction: v_off_ini = 2132.242 uV, v_off_fin = 2112.818 uV 19:30:45 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -159.818 uVs 19:30:45 Integrated Voltage Measurement... 19:30:54 Voltage Offset Correction: v_off_ini = 2134.379 uV, v_off_fin = 2148.459 uV 19:30:54 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 161.352 uVs 19:30:54 Integrated Voltage Measurement... 19:31:03 Voltage Offset Correction: v_off_ini = 2149.163 uV, v_off_fin = 2137.752 uV 19:31:03 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -177.828 uVs 19:31:03 Integrated Voltage Measurement... 19:31:12 Voltage Offset Correction: v_off_ini = 2147.548 uV, v_off_fin = 2138.117 uV 19:31:12 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 168.126 uVs 19:31:16 I1Y Average and RMS of Measurements VT = -169.265 +- 6.159 uVs I1Y = -33.853 +- 1.232 uTm 19:31:21 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 19:31:24 Integrated Voltage Measurement... 19:31:33 Voltage Offset Correction: v_off_ini = 2127.410 uV, v_off_fin = 2119.806 uV 19:31:33 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -100.988 uVs 19:31:33 Integrated Voltage Measurement... 19:31:42 Voltage Offset Correction: v_off_ini = 2128.681 uV, v_off_fin = 2132.303 uV 19:31:42 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 117.321 uVs 19:31:42 Integrated Voltage Measurement... 19:31:51 Voltage Offset Correction: v_off_ini = 2135.125 uV, v_off_fin = 2132.045 uV 19:31:51 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -99.693 uVs 19:31:51 Integrated Voltage Measurement... 19:32:00 Voltage Offset Correction: v_off_ini = 2146.526 uV, v_off_fin = 2151.292 uV 19:32:00 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 97.030 uVs 19:32:00 Integrated Voltage Measurement... 19:32:09 Voltage Offset Correction: v_off_ini = 2132.980 uV, v_off_fin = 2139.616 uV 19:32:09 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -72.009 uVs 19:32:09 Integrated Voltage Measurement... 19:32:18 Voltage Offset Correction: v_off_ini = 2135.500 uV, v_off_fin = 2167.234 uV 19:32:18 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 87.514 uVs 19:32:20 I2X Average and RMS of Measurements VT = -95.759 +- 12.140 uVs I2X = 8.810 +- 1.117 uTm^2 19:32:25 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 19:32:28 Integrated Voltage Measurement... 19:32:37 Voltage Offset Correction: v_off_ini = 2149.127 uV, v_off_fin = 2167.373 uV 19:32:37 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -187.600 uVs 19:32:37 Integrated Voltage Measurement... 19:32:46 Voltage Offset Correction: v_off_ini = 2128.472 uV, v_off_fin = 2154.933 uV 19:32:46 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 176.441 uVs 19:32:46 Integrated Voltage Measurement... 19:32:55 Voltage Offset Correction: v_off_ini = 2138.010 uV, v_off_fin = 2118.724 uV 19:32:55 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -173.967 uVs 19:32:55 Integrated Voltage Measurement... 19:33:04 Voltage Offset Correction: v_off_ini = 2110.431 uV, v_off_fin = 2098.543 uV 19:33:04 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 192.682 uVs 19:33:04 Integrated Voltage Measurement... 19:33:13 Voltage Offset Correction: v_off_ini = 2124.908 uV, v_off_fin = 2119.320 uV 19:33:13 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -183.353 uVs 19:33:13 Integrated Voltage Measurement... 19:33:22 Voltage Offset Correction: v_off_ini = 2120.934 uV, v_off_fin = 2123.668 uV 19:33:22 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 177.266 uVs 19:33:24 I2Y Average and RMS of Measurements VT = -181.885 +- 1.235 uVs I2Y = -16.733 +- 0.114 uTm^2 19:33:29 Wire Position: X0 = 1.0000 mm, Y0 = 0.0000 mm 19:33:33 Integrated Voltage Measurement... 19:33:42 Voltage Offset Correction: v_off_ini = 2131.955 uV, v_off_fin = 2129.657 uV 19:33:42 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -103.847 uVs 19:33:42 Integrated Voltage Measurement... 19:33:51 Voltage Offset Correction: v_off_ini = 2115.013 uV, v_off_fin = 2110.969 uV 19:33:51 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 81.422 uVs 19:33:51 Integrated Voltage Measurement... 19:34:00 Voltage Offset Correction: v_off_ini = 2154.049 uV, v_off_fin = 2123.730 uV 19:34:00 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -78.731 uVs 19:34:00 Integrated Voltage Measurement... 19:34:09 Voltage Offset Correction: v_off_ini = 2112.606 uV, v_off_fin = 2131.689 uV 19:34:09 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 93.651 uVs 19:34:09 Integrated Voltage Measurement... 19:34:18 Voltage Offset Correction: v_off_ini = 2122.961 uV, v_off_fin = 2113.897 uV 19:34:18 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.4999 mm dXY = -0.4999 mm VT = -65.121 uVs 19:34:18 Integrated Voltage Measurement... 19:34:27 Voltage Offset Correction: v_off_ini = 2139.153 uV, v_off_fin = 2126.790 uV 19:34:27 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = 81.568 uVs 19:34:31 I1X Average and RMS of Measurements VT = -84.057 +- 8.018 uVs I1X = 16.812 +- 1.603 uTm 19:34:34 Wire Position: X0 = 1.0000 mm, Y0 = 0.0000 mm 19:34:38 Integrated Voltage Measurement... 19:34:47 Voltage Offset Correction: v_off_ini = 2102.641 uV, v_off_fin = 2094.000 uV 19:34:47 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -189.274 uVs 19:34:47 Integrated Voltage Measurement... 19:34:56 Voltage Offset Correction: v_off_ini = 2101.239 uV, v_off_fin = 2123.231 uV 19:34:56 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 189.733 uVs 19:34:56 Integrated Voltage Measurement... 19:35:05 Voltage Offset Correction: v_off_ini = 2125.539 uV, v_off_fin = 2116.849 uV 19:35:05 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -166.501 uVs 19:35:05 Integrated Voltage Measurement... 19:35:14 Voltage Offset Correction: v_off_ini = 2118.921 uV, v_off_fin = 2128.352 uV 19:35:14 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 170.297 uVs 19:35:14 Integrated Voltage Measurement... 19:35:23 Voltage Offset Correction: v_off_ini = 2112.703 uV, v_off_fin = 2100.428 uV 19:35:23 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -184.545 uVs 19:35:23 Integrated Voltage Measurement... 19:35:32 Voltage Offset Correction: v_off_ini = 2115.002 uV, v_off_fin = 2120.042 uV 19:35:32 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 189.402 uVs 19:35:36 I1Y Average and RMS of Measurements VT = -181.625 +- 9.409 uVs I1Y = -36.325 +- 1.882 uTm 19:35:41 Wire Position: X0 = 1.0000 mm, Y0 = 0.0000 mm 19:35:44 Integrated Voltage Measurement... 19:35:53 Voltage Offset Correction: v_off_ini = 2124.765 uV, v_off_fin = 2089.042 uV 19:35:53 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -81.932 uVs 19:35:53 Integrated Voltage Measurement... 19:36:02 Voltage Offset Correction: v_off_ini = 2115.850 uV, v_off_fin = 2115.810 uV 19:36:02 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 71.335 uVs 19:36:02 Integrated Voltage Measurement... 19:36:11 Voltage Offset Correction: v_off_ini = 2108.507 uV, v_off_fin = 2125.164 uV 19:36:11 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -56.227 uVs 19:36:11 Integrated Voltage Measurement... 19:36:20 Voltage Offset Correction: v_off_ini = 2148.076 uV, v_off_fin = 2139.185 uV 19:36:20 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 65.567 uVs 19:36:20 Integrated Voltage Measurement... 19:36:29 Voltage Offset Correction: v_off_ini = 2117.759 uV, v_off_fin = 2127.093 uV 19:36:29 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -61.091 uVs 19:36:29 Integrated Voltage Measurement... 19:36:38 Voltage Offset Correction: v_off_ini = 2129.901 uV, v_off_fin = 2141.385 uV 19:36:38 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 60.974 uVs 19:36:40 I2X Average and RMS of Measurements VT = -66.188 +- 7.387 uVs I2X = 6.089 +- 0.680 uTm^2 19:36:45 Wire Position: X0 = 1.0000 mm, Y0 = 0.0000 mm 19:36:48 Integrated Voltage Measurement... 19:36:57 Voltage Offset Correction: v_off_ini = 2114.597 uV, v_off_fin = 2125.915 uV 19:36:57 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -184.444 uVs 19:36:57 Integrated Voltage Measurement... 19:37:06 Voltage Offset Correction: v_off_ini = 2127.663 uV, v_off_fin = 2130.787 uV 19:37:06 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 192.543 uVs 19:37:06 Integrated Voltage Measurement... 19:37:15 Voltage Offset Correction: v_off_ini = 2132.738 uV, v_off_fin = 2143.689 uV 19:37:15 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = -205.330 uVs 19:37:15 Integrated Voltage Measurement... 19:37:24 Voltage Offset Correction: v_off_ini = 2135.248 uV, v_off_fin = 2152.597 uV 19:37:24 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 196.254 uVs 19:37:24 Integrated Voltage Measurement... 19:37:33 Voltage Offset Correction: v_off_ini = 2157.423 uV, v_off_fin = 2137.324 uV 19:37:33 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -197.066 uVs 19:37:33 Integrated Voltage Measurement... 19:37:42 Voltage Offset Correction: v_off_ini = 2147.605 uV, v_off_fin = 2154.744 uV 19:37:42 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 199.613 uVs 19:37:44 I2Y Average and RMS of Measurements VT = -195.875 +- 5.315 uVs I2Y = -18.021 +- 0.489 uTm^2 19:37:49 Wire Position: X0 = 2.0000 mm, Y0 = 0.0000 mm 19:37:53 Integrated Voltage Measurement... 19:38:02 Voltage Offset Correction: v_off_ini = 2144.402 uV, v_off_fin = 2165.156 uV 19:38:02 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -68.562 uVs 19:38:02 Integrated Voltage Measurement... 19:38:11 Voltage Offset Correction: v_off_ini = 2165.129 uV, v_off_fin = 2163.052 uV 19:38:11 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 43.196 uVs 19:38:11 Integrated Voltage Measurement... 19:38:20 Voltage Offset Correction: v_off_ini = 2147.744 uV, v_off_fin = 2152.940 uV 19:38:20 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -75.774 uVs 19:38:20 Integrated Voltage Measurement... 19:38:29 Voltage Offset Correction: v_off_ini = 2154.230 uV, v_off_fin = 2156.291 uV 19:38:29 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 64.022 uVs 19:38:29 Integrated Voltage Measurement... 19:38:38 Voltage Offset Correction: v_off_ini = 2170.744 uV, v_off_fin = 2162.779 uV 19:38:38 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -66.097 uVs 19:38:38 Integrated Voltage Measurement... 19:38:47 Voltage Offset Correction: v_off_ini = 2170.366 uV, v_off_fin = 2152.504 uV 19:38:47 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 44.252 uVs 19:38:51 I1X Average and RMS of Measurements VT = -60.317 +- 6.781 uVs I1X = 12.063 +- 1.356 uTm 19:38:54 Wire Position: X0 = 2.0000 mm, Y0 = 0.0000 mm 19:38:58 Integrated Voltage Measurement... 19:39:07 Voltage Offset Correction: v_off_ini = 2189.543 uV, v_off_fin = 2149.292 uV 19:39:07 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -195.408 uVs 19:39:07 Integrated Voltage Measurement... 19:39:16 Voltage Offset Correction: v_off_ini = 2180.176 uV, v_off_fin = 2187.808 uV 19:39:16 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 166.288 uVs 19:39:16 Integrated Voltage Measurement... 19:39:25 Voltage Offset Correction: v_off_ini = 2170.299 uV, v_off_fin = 2167.117 uV 19:39:25 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -190.902 uVs 19:39:25 Integrated Voltage Measurement... 19:39:34 Voltage Offset Correction: v_off_ini = 2178.896 uV, v_off_fin = 2179.218 uV 19:39:34 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 198.990 uVs 19:39:34 Integrated Voltage Measurement... 19:39:43 Voltage Offset Correction: v_off_ini = 2164.756 uV, v_off_fin = 2149.263 uV 19:39:43 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -208.002 uVs 19:39:43 Integrated Voltage Measurement... 19:39:52 Voltage Offset Correction: v_off_ini = 2148.217 uV, v_off_fin = 2164.569 uV 19:39:52 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 196.104 uVs 19:39:56 I1Y Average and RMS of Measurements VT = -192.616 +- 8.812 uVs I1Y = -38.523 +- 1.762 uTm 19:40:01 Wire Position: X0 = 2.0000 mm, Y0 = 0.0000 mm 19:40:04 Integrated Voltage Measurement... 19:40:13 Voltage Offset Correction: v_off_ini = 2179.733 uV, v_off_fin = 2161.710 uV 19:40:13 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -40.124 uVs 19:40:13 Integrated Voltage Measurement... 19:40:22 Voltage Offset Correction: v_off_ini = 2156.000 uV, v_off_fin = 2170.963 uV 19:40:22 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 46.448 uVs 19:40:22 Integrated Voltage Measurement... 19:40:31 Voltage Offset Correction: v_off_ini = 2177.648 uV, v_off_fin = 2148.837 uV 19:40:31 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -32.914 uVs 19:40:31 Integrated Voltage Measurement... 19:40:40 Voltage Offset Correction: v_off_ini = 2173.607 uV, v_off_fin = 2174.671 uV 19:40:40 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 48.748 uVs 19:40:40 Integrated Voltage Measurement... 19:40:49 Voltage Offset Correction: v_off_ini = 2170.717 uV, v_off_fin = 2166.285 uV 19:40:49 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -28.791 uVs 19:40:49 Integrated Voltage Measurement... 19:40:58 Voltage Offset Correction: v_off_ini = 2192.129 uV, v_off_fin = 2185.052 uV 19:40:58 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 28.563 uVs 19:41:00 I2X Average and RMS of Measurements VT = -37.598 +- 6.387 uVs I2X = 3.459 +- 0.588 uTm^2 19:41:05 Wire Position: X0 = 2.0000 mm, Y0 = 0.0000 mm 19:41:08 Integrated Voltage Measurement... 19:41:17 Voltage Offset Correction: v_off_ini = 2174.518 uV, v_off_fin = 2186.217 uV 19:41:17 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -160.666 uVs 19:41:17 Integrated Voltage Measurement... 19:41:26 Voltage Offset Correction: v_off_ini = 2227.872 uV, v_off_fin = 2211.173 uV 19:41:26 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 156.732 uVs 19:41:26 Integrated Voltage Measurement... 19:41:35 Voltage Offset Correction: v_off_ini = 2201.344 uV, v_off_fin = 2202.954 uV 19:41:35 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -200.338 uVs 19:41:35 Integrated Voltage Measurement... 19:41:44 Voltage Offset Correction: v_off_ini = 2224.556 uV, v_off_fin = 2201.503 uV 19:41:44 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 160.293 uVs 19:41:44 Integrated Voltage Measurement... 19:41:53 Voltage Offset Correction: v_off_ini = 2202.567 uV, v_off_fin = 2198.839 uV 19:41:53 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -185.871 uVs 19:41:53 Integrated Voltage Measurement... 19:42:02 Voltage Offset Correction: v_off_ini = 2195.767 uV, v_off_fin = 2185.019 uV 19:42:02 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 202.752 uVs 19:42:04 I2Y Average and RMS of Measurements VT = -177.775 +- 14.649 uVs I2Y = -16.355 +- 1.348 uTm^2 19:42:12 Wire Position: X0 = 0.0000 mm, Y0 = -1.0000 mm 19:42:16 Integrated Voltage Measurement... 19:42:25 Voltage Offset Correction: v_off_ini = 2219.251 uV, v_off_fin = 2203.985 uV 19:42:25 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -122.524 uVs 19:42:25 Integrated Voltage Measurement... 19:42:34 Voltage Offset Correction: v_off_ini = 2215.734 uV, v_off_fin = 2205.262 uV 19:42:34 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 141.151 uVs 19:42:34 Integrated Voltage Measurement... 19:42:43 Voltage Offset Correction: v_off_ini = 2205.601 uV, v_off_fin = 2216.058 uV 19:42:43 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -119.934 uVs 19:42:43 Integrated Voltage Measurement... 19:42:52 Voltage Offset Correction: v_off_ini = 2214.953 uV, v_off_fin = 2220.005 uV 19:42:52 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 115.543 uVs 19:42:52 Integrated Voltage Measurement... 19:43:01 Voltage Offset Correction: v_off_ini = 2221.446 uV, v_off_fin = 2193.732 uV 19:43:01 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -108.861 uVs 19:43:01 Integrated Voltage Measurement... 19:43:10 Voltage Offset Correction: v_off_ini = 2191.248 uV, v_off_fin = 2195.196 uV 19:43:10 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 124.778 uVs 19:43:14 I1X Average and RMS of Measurements VT = -122.132 +- 6.873 uVs I1X = 24.427 +- 1.375 uTm 19:43:17 Wire Position: X0 = 0.0000 mm, Y0 = -1.0000 mm 19:43:21 Integrated Voltage Measurement... 19:43:30 Voltage Offset Correction: v_off_ini = 2187.414 uV, v_off_fin = 2185.496 uV 19:43:30 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -199.680 uVs 19:43:30 Integrated Voltage Measurement... 19:43:39 Voltage Offset Correction: v_off_ini = 2196.654 uV, v_off_fin = 2194.849 uV 19:43:39 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 198.821 uVs 19:43:39 Integrated Voltage Measurement... 19:43:48 Voltage Offset Correction: v_off_ini = 2179.420 uV, v_off_fin = 2164.033 uV 19:43:48 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -201.100 uVs 19:43:48 Integrated Voltage Measurement... 19:43:57 Voltage Offset Correction: v_off_ini = 2175.787 uV, v_off_fin = 2193.174 uV 19:43:57 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 199.100 uVs 19:43:57 Integrated Voltage Measurement... 19:44:06 Voltage Offset Correction: v_off_ini = 2177.528 uV, v_off_fin = 2161.185 uV 19:44:06 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -199.290 uVs 19:44:06 Integrated Voltage Measurement... 19:44:15 Voltage Offset Correction: v_off_ini = 2187.093 uV, v_off_fin = 2184.714 uV 19:44:15 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 202.913 uVs 19:44:19 I1Y Average and RMS of Measurements VT = -200.150 +- 0.757 uVs I1Y = -40.030 +- 0.151 uTm 19:44:21 Wire Position: X0 = 0.0000 mm, Y0 = -1.0000 mm 19:44:24 Integrated Voltage Measurement... 19:44:33 Voltage Offset Correction: v_off_ini = 2191.900 uV, v_off_fin = 2177.828 uV 19:44:33 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -92.922 uVs 19:44:33 Integrated Voltage Measurement... 19:44:42 Voltage Offset Correction: v_off_ini = 2196.382 uV, v_off_fin = 2208.652 uV 19:44:42 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 106.706 uVs 19:44:42 Integrated Voltage Measurement... 19:44:51 Voltage Offset Correction: v_off_ini = 2194.467 uV, v_off_fin = 2176.212 uV 19:44:51 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -78.310 uVs 19:44:51 Integrated Voltage Measurement... 19:45:00 Voltage Offset Correction: v_off_ini = 2200.837 uV, v_off_fin = 2216.343 uV 19:45:00 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 107.795 uVs 19:45:00 Integrated Voltage Measurement... 19:45:09 Voltage Offset Correction: v_off_ini = 2200.769 uV, v_off_fin = 2210.921 uV 19:45:09 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -97.671 uVs 19:45:09 Integrated Voltage Measurement... 19:45:18 Voltage Offset Correction: v_off_ini = 2226.460 uV, v_off_fin = 2235.357 uV 19:45:18 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 90.980 uVs 19:45:20 I2X Average and RMS of Measurements VT = -95.731 +- 2.934 uVs I2X = 8.807 +- 0.270 uTm^2 19:45:23 Wire Position: X0 = 0.0000 mm, Y0 = -1.0000 mm 19:45:26 Integrated Voltage Measurement... 19:45:35 Voltage Offset Correction: v_off_ini = 2250.389 uV, v_off_fin = 2237.580 uV 19:45:35 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -193.945 uVs 19:45:35 Integrated Voltage Measurement... 19:45:44 Voltage Offset Correction: v_off_ini = 2244.185 uV, v_off_fin = 2241.439 uV 19:45:44 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 180.526 uVs 19:45:44 Integrated Voltage Measurement... 19:45:53 Voltage Offset Correction: v_off_ini = 2229.494 uV, v_off_fin = 2219.255 uV 19:45:53 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -188.119 uVs 19:45:53 Integrated Voltage Measurement... 19:46:02 Voltage Offset Correction: v_off_ini = 2246.752 uV, v_off_fin = 2242.204 uV 19:46:02 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 218.321 uVs 19:46:02 Integrated Voltage Measurement... 19:46:11 Voltage Offset Correction: v_off_ini = 2221.283 uV, v_off_fin = 2233.964 uV 19:46:11 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -221.935 uVs 19:46:11 Integrated Voltage Measurement... 19:46:20 Voltage Offset Correction: v_off_ini = 2217.104 uV, v_off_fin = 2221.818 uV 19:46:20 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 203.080 uVs 19:46:22 I2Y Average and RMS of Measurements VT = -200.988 +- 10.437 uVs I2Y = -18.491 +- 0.960 uTm^2 19:46:26 Wire Position: X0 = 0.0000 mm, Y0 = -0.5000 mm 19:46:30 Integrated Voltage Measurement... 19:46:39 Voltage Offset Correction: v_off_ini = 2235.637 uV, v_off_fin = 2215.249 uV 19:46:39 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -101.223 uVs 19:46:39 Integrated Voltage Measurement... 19:46:48 Voltage Offset Correction: v_off_ini = 2207.519 uV, v_off_fin = 2245.448 uV 19:46:48 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 99.736 uVs 19:46:48 Integrated Voltage Measurement... 19:46:57 Voltage Offset Correction: v_off_ini = 2230.883 uV, v_off_fin = 2229.657 uV 19:46:57 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -127.260 uVs 19:46:57 Integrated Voltage Measurement... 19:47:06 Voltage Offset Correction: v_off_ini = 2234.077 uV, v_off_fin = 2204.628 uV 19:47:06 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 111.488 uVs 19:47:06 Integrated Voltage Measurement... 19:47:15 Voltage Offset Correction: v_off_ini = 2223.296 uV, v_off_fin = 2223.860 uV 19:47:15 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -111.794 uVs 19:47:15 Integrated Voltage Measurement... 19:47:24 Voltage Offset Correction: v_off_ini = 2219.841 uV, v_off_fin = 2201.135 uV 19:47:24 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 131.891 uVs 19:47:28 I1X Average and RMS of Measurements VT = -113.899 +- 9.542 uVs I1X = 22.780 +- 1.908 uTm 19:47:31 Wire Position: X0 = 0.0000 mm, Y0 = -0.5000 mm 19:47:35 Integrated Voltage Measurement... 19:47:44 Voltage Offset Correction: v_off_ini = 2213.461 uV, v_off_fin = 2207.818 uV 19:47:44 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -149.220 uVs 19:47:44 Integrated Voltage Measurement... 19:47:53 Voltage Offset Correction: v_off_ini = 2231.485 uV, v_off_fin = 2247.757 uV 19:47:53 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 186.126 uVs 19:47:53 Integrated Voltage Measurement... 19:48:02 Voltage Offset Correction: v_off_ini = 2228.508 uV, v_off_fin = 2215.908 uV 19:48:02 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -189.492 uVs 19:48:02 Integrated Voltage Measurement... 19:48:11 Voltage Offset Correction: v_off_ini = 2230.854 uV, v_off_fin = 2237.158 uV 19:48:11 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 178.745 uVs 19:48:11 Integrated Voltage Measurement... 19:48:20 Voltage Offset Correction: v_off_ini = 2222.980 uV, v_off_fin = 2218.917 uV 19:48:20 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -181.859 uVs 19:48:20 Integrated Voltage Measurement... 19:48:29 Voltage Offset Correction: v_off_ini = 2227.424 uV, v_off_fin = 2245.846 uV 19:48:29 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 179.267 uVs 19:48:33 I1Y Average and RMS of Measurements VT = -177.452 +- 7.065 uVs I1Y = -35.490 +- 1.413 uTm 19:48:36 Wire Position: X0 = 0.0000 mm, Y0 = -0.5000 mm 19:48:38 Integrated Voltage Measurement... 19:48:47 Voltage Offset Correction: v_off_ini = 2206.802 uV, v_off_fin = 2213.741 uV 19:48:47 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -101.178 uVs 19:48:47 Integrated Voltage Measurement... 19:48:56 Voltage Offset Correction: v_off_ini = 2217.968 uV, v_off_fin = 2236.806 uV 19:48:56 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 92.886 uVs 19:48:56 Integrated Voltage Measurement... 19:49:05 Voltage Offset Correction: v_off_ini = 2244.104 uV, v_off_fin = 2244.896 uV 19:49:05 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -81.964 uVs 19:49:05 Integrated Voltage Measurement... 19:49:14 Voltage Offset Correction: v_off_ini = 2229.770 uV, v_off_fin = 2246.672 uV 19:49:14 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 125.333 uVs 19:49:14 Integrated Voltage Measurement... 19:49:23 Voltage Offset Correction: v_off_ini = 2236.326 uV, v_off_fin = 2215.129 uV 19:49:23 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -80.358 uVs 19:49:23 Integrated Voltage Measurement... 19:49:32 Voltage Offset Correction: v_off_ini = 2217.301 uV, v_off_fin = 2214.793 uV 19:49:32 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 94.115 uVs 19:49:35 I2X Average and RMS of Measurements VT = -95.972 +- 6.742 uVs I2X = 8.829 +- 0.620 uTm^2 19:49:37 Wire Position: X0 = 0.0000 mm, Y0 = -0.5000 mm 19:49:40 Integrated Voltage Measurement... 19:49:49 Voltage Offset Correction: v_off_ini = 2194.658 uV, v_off_fin = 2199.590 uV 19:49:49 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -165.798 uVs 19:49:49 Integrated Voltage Measurement... 19:49:58 Voltage Offset Correction: v_off_ini = 2214.812 uV, v_off_fin = 2226.301 uV 19:49:58 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 191.352 uVs 19:49:58 Integrated Voltage Measurement... 19:50:07 Voltage Offset Correction: v_off_ini = 2209.198 uV, v_off_fin = 2212.354 uV 19:50:07 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -185.698 uVs 19:50:07 Integrated Voltage Measurement... 19:50:16 Voltage Offset Correction: v_off_ini = 2225.153 uV, v_off_fin = 2205.018 uV 19:50:16 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 174.684 uVs 19:50:16 Integrated Voltage Measurement... 19:50:25 Voltage Offset Correction: v_off_ini = 2210.306 uV, v_off_fin = 2203.077 uV 19:50:25 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -194.484 uVs 19:50:25 Integrated Voltage Measurement... 19:50:34 Voltage Offset Correction: v_off_ini = 2234.713 uV, v_off_fin = 2239.702 uV 19:50:34 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0001 mm dXY = -1.0000 mm VT = 200.245 uVs 19:50:36 I2Y Average and RMS of Measurements VT = -185.377 +- 8.502 uVs I2Y = -17.055 +- 0.782 uTm^2 19:50:40 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 19:50:44 Integrated Voltage Measurement... 19:50:53 Voltage Offset Correction: v_off_ini = 2215.363 uV, v_off_fin = 2256.854 uV 19:50:53 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -100.033 uVs 19:50:53 Integrated Voltage Measurement... 19:51:02 Voltage Offset Correction: v_off_ini = 2239.559 uV, v_off_fin = 2244.394 uV 19:51:02 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 101.319 uVs 19:51:02 Integrated Voltage Measurement... 19:51:11 Voltage Offset Correction: v_off_ini = 2251.000 uV, v_off_fin = 2241.560 uV 19:51:11 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -98.660 uVs 19:51:11 Integrated Voltage Measurement... 19:51:20 Voltage Offset Correction: v_off_ini = 2243.563 uV, v_off_fin = 2252.714 uV 19:51:20 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 97.648 uVs 19:51:20 Integrated Voltage Measurement... 19:51:29 Voltage Offset Correction: v_off_ini = 2250.284 uV, v_off_fin = 2256.357 uV 19:51:29 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -122.196 uVs 19:51:29 Integrated Voltage Measurement... 19:51:38 Voltage Offset Correction: v_off_ini = 2272.449 uV, v_off_fin = 2252.017 uV 19:51:38 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = 88.363 uVs 19:51:43 I1X Average and RMS of Measurements VT = -101.370 +- 2.950 uVs I1X = 20.275 +- 0.591 uTm 19:51:45 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 19:51:49 Integrated Voltage Measurement... 19:51:58 Voltage Offset Correction: v_off_ini = 2249.106 uV, v_off_fin = 2235.705 uV 19:51:58 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -177.086 uVs 19:51:58 Integrated Voltage Measurement... 19:52:07 Voltage Offset Correction: v_off_ini = 2237.905 uV, v_off_fin = 2259.783 uV 19:52:07 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 186.787 uVs 19:52:07 Integrated Voltage Measurement... 19:52:16 Voltage Offset Correction: v_off_ini = 2265.015 uV, v_off_fin = 2250.987 uV 19:52:16 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -173.497 uVs 19:52:16 Integrated Voltage Measurement... 19:52:25 Voltage Offset Correction: v_off_ini = 2254.936 uV, v_off_fin = 2246.406 uV 19:52:25 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 157.932 uVs 19:52:25 Integrated Voltage Measurement... 19:52:34 Voltage Offset Correction: v_off_ini = 2254.075 uV, v_off_fin = 2221.899 uV 19:52:34 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -183.232 uVs 19:52:34 Integrated Voltage Measurement... 19:52:43 Voltage Offset Correction: v_off_ini = 2221.240 uV, v_off_fin = 2248.236 uV 19:52:43 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 182.264 uVs 19:52:47 I1Y Average and RMS of Measurements VT = -176.800 +- 7.846 uVs I1Y = -35.360 +- 1.569 uTm 19:52:50 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 19:52:52 Integrated Voltage Measurement... 19:53:01 Voltage Offset Correction: v_off_ini = 2246.493 uV, v_off_fin = 2227.917 uV 19:53:01 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -79.817 uVs 19:53:01 Integrated Voltage Measurement... 19:53:10 Voltage Offset Correction: v_off_ini = 2238.544 uV, v_off_fin = 2272.188 uV 19:53:10 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 115.314 uVs 19:53:10 Integrated Voltage Measurement... 19:53:19 Voltage Offset Correction: v_off_ini = 2252.167 uV, v_off_fin = 2243.915 uV 19:53:19 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -73.551 uVs 19:53:19 Integrated Voltage Measurement... 19:53:28 Voltage Offset Correction: v_off_ini = 2251.546 uV, v_off_fin = 2262.475 uV 19:53:28 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 82.962 uVs 19:53:28 Integrated Voltage Measurement... 19:53:37 Voltage Offset Correction: v_off_ini = 2249.249 uV, v_off_fin = 2239.271 uV 19:53:37 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -88.030 uVs 19:53:38 Integrated Voltage Measurement... 19:53:46 Voltage Offset Correction: v_off_ini = 2231.868 uV, v_off_fin = 2254.026 uV 19:53:46 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 72.846 uVs 19:53:49 I2X Average and RMS of Measurements VT = -85.420 +- 8.634 uVs I2X = 7.859 +- 0.794 uTm^2 19:53:51 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 19:53:54 Integrated Voltage Measurement... 19:54:03 Voltage Offset Correction: v_off_ini = 2241.276 uV, v_off_fin = 2241.473 uV 19:54:03 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -166.456 uVs 19:54:03 Integrated Voltage Measurement... 19:54:12 Voltage Offset Correction: v_off_ini = 2242.831 uV, v_off_fin = 2244.793 uV 19:54:12 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = 156.633 uVs 19:54:12 Integrated Voltage Measurement... 19:54:21 Voltage Offset Correction: v_off_ini = 2244.636 uV, v_off_fin = 2248.751 uV 19:54:21 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -174.754 uVs 19:54:21 Integrated Voltage Measurement... 19:54:30 Voltage Offset Correction: v_off_ini = 2243.534 uV, v_off_fin = 2243.957 uV 19:54:30 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 188.857 uVs 19:54:30 Integrated Voltage Measurement... 19:54:39 Voltage Offset Correction: v_off_ini = 2259.087 uV, v_off_fin = 2242.952 uV 19:54:39 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -167.532 uVs 19:54:39 Integrated Voltage Measurement... 19:54:48 Voltage Offset Correction: v_off_ini = 2240.430 uV, v_off_fin = 2239.913 uV 19:54:48 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 191.927 uVs 19:54:50 I2Y Average and RMS of Measurements VT = -174.360 +- 9.101 uVs I2Y = -16.041 +- 0.837 uTm^2 19:54:54 Wire Position: X0 = 0.0000 mm, Y0 = 0.5000 mm 19:54:59 Integrated Voltage Measurement... 19:55:08 Voltage Offset Correction: v_off_ini = 2261.808 uV, v_off_fin = 2233.182 uV 19:55:08 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -117.512 uVs 19:55:08 Integrated Voltage Measurement... 19:55:17 Voltage Offset Correction: v_off_ini = 2239.001 uV, v_off_fin = 2257.489 uV 19:55:17 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 104.761 uVs 19:55:17 Integrated Voltage Measurement... 19:55:26 Voltage Offset Correction: v_off_ini = 2244.853 uV, v_off_fin = 2257.873 uV 19:55:26 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -89.765 uVs 19:55:26 Integrated Voltage Measurement... 19:55:35 Voltage Offset Correction: v_off_ini = 2246.018 uV, v_off_fin = 2257.785 uV 19:55:35 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 98.924 uVs 19:55:35 Integrated Voltage Measurement... 19:55:44 Voltage Offset Correction: v_off_ini = 2251.014 uV, v_off_fin = 2247.265 uV 19:55:44 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -117.197 uVs 19:55:44 Integrated Voltage Measurement... 19:55:53 Voltage Offset Correction: v_off_ini = 2259.320 uV, v_off_fin = 2258.589 uV 19:55:53 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = 86.747 uVs 19:55:57 I1X Average and RMS of Measurements VT = -102.484 +- 6.865 uVs I1X = 20.497 +- 1.373 uTm 19:55:59 Wire Position: X0 = 0.0000 mm, Y0 = 0.5000 mm 19:56:04 Integrated Voltage Measurement... 19:56:12 Voltage Offset Correction: v_off_ini = 2257.667 uV, v_off_fin = 2248.764 uV 19:56:12 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -171.853 uVs 19:56:12 Integrated Voltage Measurement... 19:56:21 Voltage Offset Correction: v_off_ini = 2238.461 uV, v_off_fin = 2250.514 uV 19:56:21 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 171.973 uVs 19:56:21 Integrated Voltage Measurement... 19:56:30 Voltage Offset Correction: v_off_ini = 2239.380 uV, v_off_fin = 2233.099 uV 19:56:30 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -175.446 uVs 19:56:30 Integrated Voltage Measurement... 19:56:39 Voltage Offset Correction: v_off_ini = 2244.116 uV, v_off_fin = 2252.169 uV 19:56:39 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 150.290 uVs 19:56:39 Integrated Voltage Measurement... 19:56:48 Voltage Offset Correction: v_off_ini = 2244.418 uV, v_off_fin = 2229.853 uV 19:56:48 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -167.985 uVs 19:56:48 Integrated Voltage Measurement... 19:56:57 Voltage Offset Correction: v_off_ini = 2261.641 uV, v_off_fin = 2269.747 uV 19:56:57 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 168.013 uVs 19:57:02 I1Y Average and RMS of Measurements VT = -167.593 +- 3.704 uVs I1Y = -33.518 +- 0.741 uTm 19:57:04 Wire Position: X0 = 0.0000 mm, Y0 = 0.5000 mm 19:57:07 Integrated Voltage Measurement... 19:57:16 Voltage Offset Correction: v_off_ini = 2251.087 uV, v_off_fin = 2235.361 uV 19:57:16 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -84.144 uVs 19:57:16 Integrated Voltage Measurement... 19:57:25 Voltage Offset Correction: v_off_ini = 2237.661 uV, v_off_fin = 2254.502 uV 19:57:25 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 109.216 uVs 19:57:25 Integrated Voltage Measurement... 19:57:34 Voltage Offset Correction: v_off_ini = 2230.953 uV, v_off_fin = 2226.200 uV 19:57:34 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -71.842 uVs 19:57:34 Integrated Voltage Measurement... 19:57:43 Voltage Offset Correction: v_off_ini = 2239.675 uV, v_off_fin = 2245.787 uV 19:57:43 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 100.681 uVs 19:57:43 Integrated Voltage Measurement... 19:57:52 Voltage Offset Correction: v_off_ini = 2238.553 uV, v_off_fin = 2245.660 uV 19:57:52 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -87.605 uVs 19:57:52 Integrated Voltage Measurement... 19:58:01 Voltage Offset Correction: v_off_ini = 2221.325 uV, v_off_fin = 2221.182 uV 19:58:01 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 111.311 uVs 19:58:03 I2X Average and RMS of Measurements VT = -94.133 +- 5.680 uVs I2X = 8.660 +- 0.523 uTm^2 19:58:06 Wire Position: X0 = 0.0000 mm, Y0 = 0.5000 mm 19:58:08 Integrated Voltage Measurement... 19:58:17 Voltage Offset Correction: v_off_ini = 2225.335 uV, v_off_fin = 2225.593 uV 19:58:17 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -166.111 uVs 19:58:17 Integrated Voltage Measurement... 19:58:26 Voltage Offset Correction: v_off_ini = 2230.324 uV, v_off_fin = 2213.003 uV 19:58:26 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 160.530 uVs 19:58:26 Integrated Voltage Measurement... 19:58:35 Voltage Offset Correction: v_off_ini = 2218.642 uV, v_off_fin = 2219.606 uV 19:58:35 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -173.564 uVs 19:58:35 Integrated Voltage Measurement... 19:58:44 Voltage Offset Correction: v_off_ini = 2207.246 uV, v_off_fin = 2214.070 uV 19:58:44 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 173.335 uVs 19:58:44 Integrated Voltage Measurement... 19:58:53 Voltage Offset Correction: v_off_ini = 2244.997 uV, v_off_fin = 2218.824 uV 19:58:53 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -164.182 uVs 19:58:53 Integrated Voltage Measurement... 19:59:02 Voltage Offset Correction: v_off_ini = 2204.802 uV, v_off_fin = 2222.201 uV 19:59:02 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = 159.501 uVs 19:59:05 I2Y Average and RMS of Measurements VT = -166.204 +- 5.159 uVs I2Y = -15.291 +- 0.474 uTm^2 19:59:09 Wire Position: X0 = 0.0000 mm, Y0 = 1.0000 mm 19:59:13 Integrated Voltage Measurement... 19:59:22 Voltage Offset Correction: v_off_ini = 2250.791 uV, v_off_fin = 2243.393 uV 19:59:22 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -111.309 uVs 19:59:22 Integrated Voltage Measurement... 19:59:31 Voltage Offset Correction: v_off_ini = 2223.667 uV, v_off_fin = 2233.999 uV 19:59:31 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 78.872 uVs 19:59:31 Integrated Voltage Measurement... 19:59:40 Voltage Offset Correction: v_off_ini = 2245.601 uV, v_off_fin = 2246.518 uV 19:59:40 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -87.722 uVs 19:59:40 Integrated Voltage Measurement... 19:59:49 Voltage Offset Correction: v_off_ini = 2230.725 uV, v_off_fin = 2244.666 uV 19:59:49 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 103.989 uVs 19:59:49 Integrated Voltage Measurement... 19:59:58 Voltage Offset Correction: v_off_ini = 2233.301 uV, v_off_fin = 2235.531 uV 19:59:58 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -76.202 uVs 19:59:58 Integrated Voltage Measurement... 20:00:07 Voltage Offset Correction: v_off_ini = 2236.930 uV, v_off_fin = 2243.434 uV 20:00:07 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 97.991 uVs 20:00:11 I1X Average and RMS of Measurements VT = -92.681 +- 3.961 uVs I1X = 18.536 +- 0.792 uTm 20:00:14 Wire Position: X0 = 0.0000 mm, Y0 = 1.0000 mm 20:00:18 Integrated Voltage Measurement... 20:00:27 Voltage Offset Correction: v_off_ini = 2221.612 uV, v_off_fin = 2226.901 uV 20:00:27 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -150.109 uVs 20:00:27 Integrated Voltage Measurement... 20:00:36 Voltage Offset Correction: v_off_ini = 2216.967 uV, v_off_fin = 2236.719 uV 20:00:36 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 164.994 uVs 20:00:36 Integrated Voltage Measurement... 20:00:45 Voltage Offset Correction: v_off_ini = 2231.019 uV, v_off_fin = 2213.844 uV 20:00:45 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -130.526 uVs 20:00:45 Integrated Voltage Measurement... 20:00:54 Voltage Offset Correction: v_off_ini = 2234.081 uV, v_off_fin = 2239.054 uV 20:00:54 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 151.111 uVs 20:00:54 Integrated Voltage Measurement... 20:01:03 Voltage Offset Correction: v_off_ini = 2241.857 uV, v_off_fin = 2230.880 uV 20:01:03 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -135.086 uVs 20:01:03 Integrated Voltage Measurement... 20:01:12 Voltage Offset Correction: v_off_ini = 2224.091 uV, v_off_fin = 2233.160 uV 20:01:12 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 171.388 uVs 20:01:16 I1Y Average and RMS of Measurements VT = -150.536 +- 7.093 uVs I1Y = -30.105 +- 1.418 uTm 20:01:18 Wire Position: X0 = 0.0000 mm, Y0 = 1.0000 mm 20:01:21 Integrated Voltage Measurement... 20:01:30 Voltage Offset Correction: v_off_ini = 2220.868 uV, v_off_fin = 2227.023 uV 20:01:30 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -78.868 uVs 20:01:30 Integrated Voltage Measurement... 20:01:39 Voltage Offset Correction: v_off_ini = 2236.488 uV, v_off_fin = 2238.462 uV 20:01:39 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 108.486 uVs 20:01:39 Integrated Voltage Measurement... 20:01:48 Voltage Offset Correction: v_off_ini = 2226.361 uV, v_off_fin = 2228.115 uV 20:01:48 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -67.117 uVs 20:01:48 Integrated Voltage Measurement... 20:01:57 Voltage Offset Correction: v_off_ini = 2211.781 uV, v_off_fin = 2242.292 uV 20:01:57 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 107.084 uVs 20:01:57 Integrated Voltage Measurement... 20:02:06 Voltage Offset Correction: v_off_ini = 2232.235 uV, v_off_fin = 2227.285 uV 20:02:06 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -78.102 uVs 20:02:06 Integrated Voltage Measurement... 20:02:15 Voltage Offset Correction: v_off_ini = 2243.661 uV, v_off_fin = 2280.051 uV 20:02:15 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 75.067 uVs 20:02:17 I2X Average and RMS of Measurements VT = -85.787 +- 7.040 uVs I2X = 7.892 +- 0.648 uTm^2 20:02:20 Wire Position: X0 = 0.0000 mm, Y0 = 1.0000 mm 20:02:22 Integrated Voltage Measurement... 20:02:31 Voltage Offset Correction: v_off_ini = 2248.358 uV, v_off_fin = 2273.803 uV 20:02:31 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -161.282 uVs 20:02:31 Integrated Voltage Measurement... 20:02:40 Voltage Offset Correction: v_off_ini = 2272.112 uV, v_off_fin = 2266.449 uV 20:02:40 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 137.731 uVs 20:02:40 Integrated Voltage Measurement... 20:02:49 Voltage Offset Correction: v_off_ini = 2281.918 uV, v_off_fin = 2268.384 uV 20:02:49 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -160.859 uVs 20:02:49 Integrated Voltage Measurement... 20:02:58 Voltage Offset Correction: v_off_ini = 2267.057 uV, v_off_fin = 2262.080 uV 20:02:58 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 151.395 uVs 20:02:58 Integrated Voltage Measurement... 20:03:07 Voltage Offset Correction: v_off_ini = 2264.158 uV, v_off_fin = 2261.726 uV 20:03:07 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -148.006 uVs 20:03:08 Integrated Voltage Measurement... 20:03:16 Voltage Offset Correction: v_off_ini = 2273.103 uV, v_off_fin = 2273.285 uV 20:03:16 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 143.577 uVs 20:03:19 I2Y Average and RMS of Measurements VT = -150.475 +- 4.275 uVs I2Y = -13.844 +- 0.393 uTm^2 20:03:36 Wire Position: X0 = -2.0000 mm, Y0 = 0.0000 mm 20:03:40 Integrated Voltage Measurement... 20:03:49 Voltage Offset Correction: v_off_ini = 2291.688 uV, v_off_fin = 2273.839 uV 20:03:49 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -145.680 uVs 20:03:49 Integrated Voltage Measurement... 20:03:58 Voltage Offset Correction: v_off_ini = 2274.905 uV, v_off_fin = 2293.138 uV 20:03:58 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 141.164 uVs 20:03:58 Integrated Voltage Measurement... 20:04:07 Voltage Offset Correction: v_off_ini = 2299.784 uV, v_off_fin = 2278.938 uV 20:04:07 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -137.442 uVs 20:04:07 Integrated Voltage Measurement... 20:04:16 Voltage Offset Correction: v_off_ini = 2282.606 uV, v_off_fin = 2282.873 uV 20:04:16 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = 149.975 uVs 20:04:16 Integrated Voltage Measurement... 20:04:25 Voltage Offset Correction: v_off_ini = 2269.154 uV, v_off_fin = 2285.459 uV 20:04:25 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -152.343 uVs 20:04:25 Integrated Voltage Measurement... 20:04:34 Voltage Offset Correction: v_off_ini = 2294.639 uV, v_off_fin = 2278.844 uV 20:04:34 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 156.080 uVs 20:04:38 I1X Average and RMS of Measurements VT = -147.114 +- 5.020 uVs I1X = 29.424 +- 1.003 uTm 20:04:41 Wire Position: X0 = -2.0000 mm, Y0 = 0.0000 mm 20:04:45 Integrated Voltage Measurement... 20:04:54 Voltage Offset Correction: v_off_ini = 2272.612 uV, v_off_fin = 2264.253 uV 20:04:54 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -146.015 uVs 20:04:54 Integrated Voltage Measurement... 20:05:03 Voltage Offset Correction: v_off_ini = 2269.438 uV, v_off_fin = 2275.942 uV 20:05:03 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 139.514 uVs 20:05:03 Integrated Voltage Measurement... 20:05:12 Voltage Offset Correction: v_off_ini = 2259.016 uV, v_off_fin = 2261.239 uV 20:05:12 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -140.848 uVs 20:05:12 Integrated Voltage Measurement... 20:05:21 Voltage Offset Correction: v_off_ini = 2268.096 uV, v_off_fin = 2292.659 uV 20:05:21 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 140.538 uVs 20:05:21 Integrated Voltage Measurement... 20:05:30 Voltage Offset Correction: v_off_ini = 2285.251 uV, v_off_fin = 2270.236 uV 20:05:30 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -147.491 uVs 20:05:30 Integrated Voltage Measurement... 20:05:39 Voltage Offset Correction: v_off_ini = 2283.865 uV, v_off_fin = 2288.687 uV 20:05:39 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 151.366 uVs 20:05:43 I1Y Average and RMS of Measurements VT = -144.296 +- 3.727 uVs I1Y = -28.859 +- 0.745 uTm 20:05:48 Wire Position: X0 = -2.0000 mm, Y0 = 0.0000 mm 20:05:51 Integrated Voltage Measurement... 20:06:00 Voltage Offset Correction: v_off_ini = 2274.671 uV, v_off_fin = 2261.754 uV 20:06:00 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -114.211 uVs 20:06:00 Integrated Voltage Measurement... 20:06:09 Voltage Offset Correction: v_off_ini = 2264.448 uV, v_off_fin = 2276.714 uV 20:06:09 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 135.483 uVs 20:06:09 Integrated Voltage Measurement... 20:06:18 Voltage Offset Correction: v_off_ini = 2258.317 uV, v_off_fin = 2270.500 uV 20:06:18 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -120.003 uVs 20:06:18 Integrated Voltage Measurement... 20:06:27 Voltage Offset Correction: v_off_ini = 2265.382 uV, v_off_fin = 2274.048 uV 20:06:27 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 147.080 uVs 20:06:27 Integrated Voltage Measurement... 20:06:36 Voltage Offset Correction: v_off_ini = 2261.944 uV, v_off_fin = 2255.805 uV 20:06:36 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -137.520 uVs 20:06:36 Integrated Voltage Measurement... 20:06:45 Voltage Offset Correction: v_off_ini = 2291.608 uV, v_off_fin = 2300.792 uV 20:06:45 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 128.309 uVs 20:06:47 I2X Average and RMS of Measurements VT = -130.435 +- 3.959 uVs I2X = 12.000 +- 0.364 uTm^2 20:06:52 Wire Position: X0 = -2.0000 mm, Y0 = 0.0000 mm 20:06:55 Integrated Voltage Measurement... 20:07:04 Voltage Offset Correction: v_off_ini = 2286.280 uV, v_off_fin = 2277.388 uV 20:07:04 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -118.399 uVs 20:07:04 Integrated Voltage Measurement... 20:07:13 Voltage Offset Correction: v_off_ini = 2275.284 uV, v_off_fin = 2292.005 uV 20:07:13 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 138.269 uVs 20:07:13 Integrated Voltage Measurement... 20:07:22 Voltage Offset Correction: v_off_ini = 2302.213 uV, v_off_fin = 2298.075 uV 20:07:22 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = -138.544 uVs 20:07:22 Integrated Voltage Measurement... 20:07:31 Voltage Offset Correction: v_off_ini = 2305.068 uV, v_off_fin = 2319.186 uV 20:07:31 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 151.945 uVs 20:07:31 Integrated Voltage Measurement... 20:07:40 Voltage Offset Correction: v_off_ini = 2310.568 uV, v_off_fin = 2318.085 uV 20:07:40 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -146.402 uVs 20:07:40 Integrated Voltage Measurement... 20:07:49 Voltage Offset Correction: v_off_ini = 2306.039 uV, v_off_fin = 2303.318 uV 20:07:49 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 157.197 uVs 20:07:51 I2Y Average and RMS of Measurements VT = -141.793 +- 9.886 uVs I2Y = -13.045 +- 0.910 uTm^2 20:07:56 Wire Position: X0 = -1.0000 mm, Y0 = 0.0000 mm 20:08:00 Integrated Voltage Measurement... 20:08:09 Voltage Offset Correction: v_off_ini = 2282.687 uV, v_off_fin = 2301.781 uV 20:08:09 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -116.454 uVs 20:08:09 Integrated Voltage Measurement... 20:08:18 Voltage Offset Correction: v_off_ini = 2300.037 uV, v_off_fin = 2295.960 uV 20:08:18 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 125.253 uVs 20:08:18 Integrated Voltage Measurement... 20:08:27 Voltage Offset Correction: v_off_ini = 2270.367 uV, v_off_fin = 2301.279 uV 20:08:27 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -118.654 uVs 20:08:27 Integrated Voltage Measurement... 20:08:36 Voltage Offset Correction: v_off_ini = 2309.983 uV, v_off_fin = 2312.385 uV 20:08:36 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 145.336 uVs 20:08:36 Integrated Voltage Measurement... 20:08:45 Voltage Offset Correction: v_off_ini = 2314.037 uV, v_off_fin = 2306.278 uV 20:08:45 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -140.769 uVs 20:08:45 Integrated Voltage Measurement... 20:08:54 Voltage Offset Correction: v_off_ini = 2325.935 uV, v_off_fin = 2319.780 uV 20:08:54 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 92.067 uVs 20:08:58 I1X Average and RMS of Measurements VT = -123.089 +- 6.553 uVs I1X = 24.619 +- 1.311 uTm 20:09:01 Wire Position: X0 = -1.0000 mm, Y0 = 0.0000 mm 20:09:05 Integrated Voltage Measurement... 20:09:14 Voltage Offset Correction: v_off_ini = 2330.524 uV, v_off_fin = 2333.081 uV 20:09:14 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -142.576 uVs 20:09:14 Integrated Voltage Measurement... 20:09:23 Voltage Offset Correction: v_off_ini = 2341.969 uV, v_off_fin = 2329.763 uV 20:09:23 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 165.052 uVs 20:09:23 Integrated Voltage Measurement... 20:09:32 Voltage Offset Correction: v_off_ini = 2322.807 uV, v_off_fin = 2314.591 uV 20:09:32 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -151.595 uVs 20:09:32 Integrated Voltage Measurement... 20:09:41 Voltage Offset Correction: v_off_ini = 2322.589 uV, v_off_fin = 2322.284 uV 20:09:41 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 163.670 uVs 20:09:41 Integrated Voltage Measurement... 20:09:50 Voltage Offset Correction: v_off_ini = 2338.321 uV, v_off_fin = 2318.612 uV 20:09:50 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -164.188 uVs 20:09:50 Integrated Voltage Measurement... 20:09:59 Voltage Offset Correction: v_off_ini = 2306.418 uV, v_off_fin = 2309.971 uV 20:09:59 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 156.941 uVs 20:10:03 I1Y Average and RMS of Measurements VT = -157.337 +- 2.764 uVs I1Y = -31.468 +- 0.552 uTm 20:10:08 Wire Position: X0 = -1.0000 mm, Y0 = 0.0000 mm 20:10:11 Integrated Voltage Measurement... 20:10:20 Voltage Offset Correction: v_off_ini = 2308.665 uV, v_off_fin = 2303.968 uV 20:10:20 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -121.425 uVs 20:10:20 Integrated Voltage Measurement... 20:10:29 Voltage Offset Correction: v_off_ini = 2326.362 uV, v_off_fin = 2326.602 uV 20:10:29 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 104.779 uVs 20:10:29 Integrated Voltage Measurement... 20:10:38 Voltage Offset Correction: v_off_ini = 2290.661 uV, v_off_fin = 2306.276 uV 20:10:38 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -110.502 uVs 20:10:38 Integrated Voltage Measurement... 20:10:47 Voltage Offset Correction: v_off_ini = 2322.054 uV, v_off_fin = 2322.081 uV 20:10:47 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 118.656 uVs 20:10:47 Integrated Voltage Measurement... 20:10:56 Voltage Offset Correction: v_off_ini = 2301.545 uV, v_off_fin = 2297.271 uV 20:10:56 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -77.576 uVs 20:10:56 Integrated Voltage Measurement... 20:11:05 Voltage Offset Correction: v_off_ini = 2323.855 uV, v_off_fin = 2307.921 uV 20:11:05 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 102.598 uVs 20:11:07 I2X Average and RMS of Measurements VT = -105.923 +- 11.214 uVs I2X = 9.745 +- 1.032 uTm^2 20:11:12 Wire Position: X0 = -1.0000 mm, Y0 = 0.0000 mm 20:11:15 Integrated Voltage Measurement... 20:11:24 Voltage Offset Correction: v_off_ini = 2320.257 uV, v_off_fin = 2326.397 uV 20:11:24 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -158.316 uVs 20:11:24 Integrated Voltage Measurement... 20:11:33 Voltage Offset Correction: v_off_ini = 2321.823 uV, v_off_fin = 2336.252 uV 20:11:33 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 137.640 uVs 20:11:33 Integrated Voltage Measurement... 20:11:42 Voltage Offset Correction: v_off_ini = 2327.995 uV, v_off_fin = 2300.841 uV 20:11:42 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -139.323 uVs 20:11:42 Integrated Voltage Measurement... 20:11:51 Voltage Offset Correction: v_off_ini = 2313.728 uV, v_off_fin = 2347.460 uV 20:11:51 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 157.408 uVs 20:11:51 Integrated Voltage Measurement... 20:12:00 Voltage Offset Correction: v_off_ini = 2312.899 uV, v_off_fin = 2319.064 uV 20:12:00 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -149.362 uVs 20:12:00 Integrated Voltage Measurement... 20:12:09 Voltage Offset Correction: v_off_ini = 2350.109 uV, v_off_fin = 2344.851 uV 20:12:09 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 150.565 uVs 20:12:11 I2Y Average and RMS of Measurements VT = -148.769 +- 0.859 uVs I2Y = -13.687 +- 0.079 uTm^2 20:12:16 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 20:12:20 Integrated Voltage Measurement... 20:12:29 Voltage Offset Correction: v_off_ini = 2347.639 uV, v_off_fin = 2362.649 uV 20:12:29 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -125.048 uVs 20:12:29 Integrated Voltage Measurement... 20:12:38 Voltage Offset Correction: v_off_ini = 2380.678 uV, v_off_fin = 2366.524 uV 20:12:38 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 100.017 uVs 20:12:38 Integrated Voltage Measurement... 20:12:47 Voltage Offset Correction: v_off_ini = 2354.469 uV, v_off_fin = 2345.330 uV 20:12:47 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -117.222 uVs 20:12:47 Integrated Voltage Measurement... 20:12:56 Voltage Offset Correction: v_off_ini = 2346.433 uV, v_off_fin = 2374.924 uV 20:12:56 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 110.789 uVs 20:12:56 Integrated Voltage Measurement... 20:13:05 Voltage Offset Correction: v_off_ini = 2379.380 uV, v_off_fin = 2355.240 uV 20:13:05 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -128.439 uVs 20:13:05 Integrated Voltage Measurement... 20:13:14 Voltage Offset Correction: v_off_ini = 2383.385 uV, v_off_fin = 2377.197 uV 20:13:14 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 92.514 uVs 20:13:18 I1X Average and RMS of Measurements VT = -112.338 +- 1.447 uVs I1X = 22.468 +- 0.289 uTm 20:13:21 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 20:13:25 Integrated Voltage Measurement... 20:13:34 Voltage Offset Correction: v_off_ini = 2400.742 uV, v_off_fin = 2370.761 uV 20:13:34 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -148.949 uVs 20:13:34 Integrated Voltage Measurement... 20:13:43 Voltage Offset Correction: v_off_ini = 2381.363 uV, v_off_fin = 2402.791 uV 20:13:43 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 166.393 uVs 20:13:43 Integrated Voltage Measurement... 20:13:52 Voltage Offset Correction: v_off_ini = 2396.804 uV, v_off_fin = 2386.930 uV 20:13:52 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -151.292 uVs 20:13:52 Integrated Voltage Measurement... 20:14:01 Voltage Offset Correction: v_off_ini = 2376.137 uV, v_off_fin = 2392.292 uV 20:14:01 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 159.208 uVs 20:14:01 Integrated Voltage Measurement... 20:14:10 Voltage Offset Correction: v_off_ini = 2382.765 uV, v_off_fin = 2385.170 uV 20:14:10 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -156.493 uVs 20:14:10 Integrated Voltage Measurement... 20:14:19 Voltage Offset Correction: v_off_ini = 2382.579 uV, v_off_fin = 2388.640 uV 20:14:19 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 154.102 uVs 20:14:23 I1Y Average and RMS of Measurements VT = -156.073 +- 1.130 uVs I1Y = -31.215 +- 0.226 uTm 20:14:28 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 20:14:31 Integrated Voltage Measurement... 20:14:40 Voltage Offset Correction: v_off_ini = 2367.282 uV, v_off_fin = 2374.369 uV 20:14:40 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -68.433 uVs 20:14:40 Integrated Voltage Measurement... 20:14:49 Voltage Offset Correction: v_off_ini = 2385.575 uV, v_off_fin = 2398.756 uV 20:14:49 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 88.930 uVs 20:14:49 Integrated Voltage Measurement... 20:14:58 Voltage Offset Correction: v_off_ini = 2377.549 uV, v_off_fin = 2375.071 uV 20:14:58 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -86.481 uVs 20:14:58 Integrated Voltage Measurement... 20:15:07 Voltage Offset Correction: v_off_ini = 2386.550 uV, v_off_fin = 2410.347 uV 20:15:07 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 82.264 uVs 20:15:07 Integrated Voltage Measurement... 20:15:16 Voltage Offset Correction: v_off_ini = 2403.250 uV, v_off_fin = 2394.201 uV 20:15:16 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -81.377 uVs 20:15:16 Integrated Voltage Measurement... 20:15:25 Voltage Offset Correction: v_off_ini = 2389.302 uV, v_off_fin = 2391.559 uV 20:15:25 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 79.133 uVs 20:15:27 I2X Average and RMS of Measurements VT = -81.103 +- 2.400 uVs I2X = 7.461 +- 0.221 uTm^2 20:15:32 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 20:15:35 Integrated Voltage Measurement... 20:15:44 Voltage Offset Correction: v_off_ini = 2402.033 uV, v_off_fin = 2396.891 uV 20:15:44 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -156.442 uVs 20:15:44 Integrated Voltage Measurement... 20:15:53 Voltage Offset Correction: v_off_ini = 2396.271 uV, v_off_fin = 2397.838 uV 20:15:53 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 176.771 uVs 20:15:53 Integrated Voltage Measurement... 20:16:02 Voltage Offset Correction: v_off_ini = 2396.224 uV, v_off_fin = 2412.257 uV 20:16:02 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -171.565 uVs 20:16:02 Integrated Voltage Measurement... 20:16:11 Voltage Offset Correction: v_off_ini = 2412.876 uV, v_off_fin = 2398.516 uV 20:16:11 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 149.875 uVs 20:16:11 Integrated Voltage Measurement... 20:16:20 Voltage Offset Correction: v_off_ini = 2390.499 uV, v_off_fin = 2399.189 uV 20:16:20 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -161.526 uVs 20:16:20 Integrated Voltage Measurement... 20:16:29 Voltage Offset Correction: v_off_ini = 2404.471 uV, v_off_fin = 2386.933 uV 20:16:29 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 161.691 uVs 20:16:31 I2Y Average and RMS of Measurements VT = -162.978 +- 2.591 uVs I2Y = -14.994 +- 0.238 uTm^2 20:16:36 Wire Position: X0 = 1.0000 mm, Y0 = 0.0000 mm 20:16:40 Integrated Voltage Measurement... 20:16:49 Voltage Offset Correction: v_off_ini = 2411.826 uV, v_off_fin = 2401.773 uV 20:16:49 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.4998 mm dXY = -0.4999 mm VT = -86.413 uVs 20:16:49 Integrated Voltage Measurement... 20:16:58 Voltage Offset Correction: v_off_ini = 2409.875 uV, v_off_fin = 2402.422 uV 20:16:58 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 72.504 uVs 20:16:58 Integrated Voltage Measurement... 20:17:07 Voltage Offset Correction: v_off_ini = 2399.916 uV, v_off_fin = 2382.180 uV 20:17:07 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -82.680 uVs 20:17:07 Integrated Voltage Measurement... 20:17:16 Voltage Offset Correction: v_off_ini = 2389.273 uV, v_off_fin = 2418.575 uV 20:17:16 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 62.161 uVs 20:17:16 Integrated Voltage Measurement... 20:17:25 Voltage Offset Correction: v_off_ini = 2393.319 uV, v_off_fin = 2383.375 uV 20:17:25 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -69.609 uVs 20:17:25 Integrated Voltage Measurement... 20:17:34 Voltage Offset Correction: v_off_ini = 2383.853 uV, v_off_fin = 2378.858 uV 20:17:34 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 70.005 uVs 20:17:38 I1X Average and RMS of Measurements VT = -73.895 +- 4.076 uVs I1X = 14.780 +- 0.816 uTm 20:17:41 Wire Position: X0 = 1.0000 mm, Y0 = 0.0000 mm 20:17:45 Integrated Voltage Measurement... 20:17:54 Voltage Offset Correction: v_off_ini = 2371.558 uV, v_off_fin = 2378.912 uV 20:17:54 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -165.987 uVs 20:17:54 Integrated Voltage Measurement... 20:18:03 Voltage Offset Correction: v_off_ini = 2393.193 uV, v_off_fin = 2385.581 uV 20:18:03 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 186.661 uVs 20:18:03 Integrated Voltage Measurement... 20:18:12 Voltage Offset Correction: v_off_ini = 2368.965 uV, v_off_fin = 2353.692 uV 20:18:12 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -182.476 uVs 20:18:12 Integrated Voltage Measurement... 20:18:21 Voltage Offset Correction: v_off_ini = 2365.424 uV, v_off_fin = 2369.504 uV 20:18:21 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 184.324 uVs 20:18:21 Integrated Voltage Measurement... 20:18:30 Voltage Offset Correction: v_off_ini = 2365.256 uV, v_off_fin = 2366.160 uV 20:18:30 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -164.665 uVs 20:18:30 Integrated Voltage Measurement... 20:18:39 Voltage Offset Correction: v_off_ini = 2366.012 uV, v_off_fin = 2372.616 uV 20:18:39 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 172.970 uVs 20:18:43 I1Y Average and RMS of Measurements VT = -176.181 +- 5.954 uVs I1Y = -35.236 +- 1.191 uTm 20:18:48 Wire Position: X0 = 1.0000 mm, Y0 = 0.0000 mm 20:18:50 Integrated Voltage Measurement... 20:18:59 Voltage Offset Correction: v_off_ini = 2348.454 uV, v_off_fin = 2336.814 uV 20:18:59 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -70.459 uVs 20:18:59 Integrated Voltage Measurement... 20:19:08 Voltage Offset Correction: v_off_ini = 2321.288 uV, v_off_fin = 2335.403 uV 20:19:08 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 59.881 uVs 20:19:08 Integrated Voltage Measurement... 20:19:17 Voltage Offset Correction: v_off_ini = 2334.560 uV, v_off_fin = 2335.394 uV 20:19:17 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -74.289 uVs 20:19:18 Integrated Voltage Measurement... 20:19:26 Voltage Offset Correction: v_off_ini = 2328.453 uV, v_off_fin = 2322.876 uV 20:19:26 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 59.183 uVs 20:19:27 Integrated Voltage Measurement... 20:19:35 Voltage Offset Correction: v_off_ini = 2339.168 uV, v_off_fin = 2329.298 uV 20:19:36 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -77.976 uVs 20:19:36 Integrated Voltage Measurement... 20:19:44 Voltage Offset Correction: v_off_ini = 2330.925 uV, v_off_fin = 2336.009 uV 20:19:45 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 69.604 uVs 20:19:47 I2X Average and RMS of Measurements VT = -68.565 +- 3.749 uVs I2X = 6.308 +- 0.345 uTm^2 20:19:52 Wire Position: X0 = 1.0000 mm, Y0 = 0.0000 mm 20:19:55 Integrated Voltage Measurement... 20:20:04 Voltage Offset Correction: v_off_ini = 2363.580 uV, v_off_fin = 2356.772 uV 20:20:04 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -148.481 uVs 20:20:04 Integrated Voltage Measurement... 20:20:13 Voltage Offset Correction: v_off_ini = 2350.935 uV, v_off_fin = 2364.280 uV 20:20:13 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 164.943 uVs 20:20:13 Integrated Voltage Measurement... 20:20:22 Voltage Offset Correction: v_off_ini = 2370.580 uV, v_off_fin = 2348.658 uV 20:20:22 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -156.569 uVs 20:20:22 Integrated Voltage Measurement... 20:20:31 Voltage Offset Correction: v_off_ini = 2347.199 uV, v_off_fin = 2357.574 uV 20:20:31 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 174.901 uVs 20:20:31 Integrated Voltage Measurement... 20:20:40 Voltage Offset Correction: v_off_ini = 2373.089 uV, v_off_fin = 2370.042 uV 20:20:40 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -177.071 uVs 20:20:40 Integrated Voltage Measurement... 20:20:49 Voltage Offset Correction: v_off_ini = 2375.517 uV, v_off_fin = 2374.051 uV 20:20:49 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 147.801 uVs 20:20:51 I2Y Average and RMS of Measurements VT = -161.628 +- 3.728 uVs I2Y = -14.870 +- 0.343 uTm^2 20:20:56 Wire Position: X0 = 2.0000 mm, Y0 = 0.0000 mm 20:21:00 Integrated Voltage Measurement... 20:21:09 Voltage Offset Correction: v_off_ini = 2375.301 uV, v_off_fin = 2393.149 uV 20:21:09 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -44.467 uVs 20:21:09 Integrated Voltage Measurement... 20:21:18 Voltage Offset Correction: v_off_ini = 2390.736 uV, v_off_fin = 2392.253 uV 20:21:18 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = 52.188 uVs 20:21:18 Integrated Voltage Measurement... 20:21:27 Voltage Offset Correction: v_off_ini = 2382.189 uV, v_off_fin = 2370.217 uV 20:21:27 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -49.608 uVs 20:21:27 Integrated Voltage Measurement... 20:21:36 Voltage Offset Correction: v_off_ini = 2391.000 uV, v_off_fin = 2423.762 uV 20:21:36 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 48.495 uVs 20:21:36 Integrated Voltage Measurement... 20:21:45 Voltage Offset Correction: v_off_ini = 2420.786 uV, v_off_fin = 2393.104 uV 20:21:45 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -69.700 uVs 20:21:45 Integrated Voltage Measurement... 20:21:54 Voltage Offset Correction: v_off_ini = 2415.544 uV, v_off_fin = 2404.955 uV 20:21:54 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 59.228 uVs 20:21:58 I1X Average and RMS of Measurements VT = -53.948 +- 7.442 uVs I1X = 10.790 +- 1.488 uTm 20:22:00 Wire Position: X0 = 2.0000 mm, Y0 = 0.0000 mm 20:22:05 Integrated Voltage Measurement... 20:22:14 Voltage Offset Correction: v_off_ini = 2405.429 uV, v_off_fin = 2404.294 uV 20:22:14 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -180.981 uVs 20:22:14 Integrated Voltage Measurement... 20:22:23 Voltage Offset Correction: v_off_ini = 2407.082 uV, v_off_fin = 2419.364 uV 20:22:23 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 185.934 uVs 20:22:23 Integrated Voltage Measurement... 20:22:32 Voltage Offset Correction: v_off_ini = 2398.076 uV, v_off_fin = 2398.872 uV 20:22:32 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -169.607 uVs 20:22:32 Integrated Voltage Measurement... 20:22:41 Voltage Offset Correction: v_off_ini = 2391.110 uV, v_off_fin = 2413.841 uV 20:22:41 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 199.119 uVs 20:22:41 Integrated Voltage Measurement... 20:22:50 Voltage Offset Correction: v_off_ini = 2403.425 uV, v_off_fin = 2385.779 uV 20:22:50 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -182.438 uVs 20:22:50 Integrated Voltage Measurement... 20:22:59 Voltage Offset Correction: v_off_ini = 2399.534 uV, v_off_fin = 2410.765 uV 20:22:59 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 180.314 uVs 20:23:03 I1Y Average and RMS of Measurements VT = -183.065 +- 1.250 uVs I1Y = -36.613 +- 0.250 uTm 20:23:08 Wire Position: X0 = 2.0000 mm, Y0 = 0.0000 mm 20:23:10 Integrated Voltage Measurement... 20:23:19 Voltage Offset Correction: v_off_ini = 2381.260 uV, v_off_fin = 2366.369 uV 20:23:19 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -35.052 uVs 20:23:19 Integrated Voltage Measurement... 20:23:28 Voltage Offset Correction: v_off_ini = 2370.294 uV, v_off_fin = 2378.512 uV 20:23:28 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 48.266 uVs 20:23:28 Integrated Voltage Measurement... 20:23:37 Voltage Offset Correction: v_off_ini = 2376.226 uV, v_off_fin = 2367.460 uV 20:23:37 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -31.006 uVs 20:23:37 Integrated Voltage Measurement... 20:23:46 Voltage Offset Correction: v_off_ini = 2387.833 uV, v_off_fin = 2394.396 uV 20:23:46 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 16.276 uVs 20:23:46 Integrated Voltage Measurement... 20:23:55 Voltage Offset Correction: v_off_ini = 2380.058 uV, v_off_fin = 2376.076 uV 20:23:55 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -29.065 uVs 20:23:55 Integrated Voltage Measurement... 20:24:04 Voltage Offset Correction: v_off_ini = 2382.369 uV, v_off_fin = 2408.018 uV 20:24:04 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 24.761 uVs 20:24:07 I2X Average and RMS of Measurements VT = -30.738 +- 7.837 uVs I2X = 2.828 +- 0.721 uTm^2 20:24:12 Wire Position: X0 = 2.0000 mm, Y0 = 0.0000 mm 20:24:14 Integrated Voltage Measurement... 20:24:23 Voltage Offset Correction: v_off_ini = 2381.223 uV, v_off_fin = 2395.513 uV 20:24:23 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -176.130 uVs 20:24:24 Integrated Voltage Measurement... 20:24:32 Voltage Offset Correction: v_off_ini = 2422.123 uV, v_off_fin = 2425.252 uV 20:24:33 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 162.233 uVs 20:24:33 Integrated Voltage Measurement... 20:24:42 Voltage Offset Correction: v_off_ini = 2407.982 uV, v_off_fin = 2411.451 uV 20:24:42 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -173.303 uVs 20:24:42 Integrated Voltage Measurement... 20:24:51 Voltage Offset Correction: v_off_ini = 2427.312 uV, v_off_fin = 2424.260 uV 20:24:51 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 149.133 uVs 20:24:51 Integrated Voltage Measurement... 20:25:00 Voltage Offset Correction: v_off_ini = 2419.964 uV, v_off_fin = 2406.243 uV 20:25:00 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -182.349 uVs 20:25:00 Integrated Voltage Measurement... 20:25:09 Voltage Offset Correction: v_off_ini = 2406.975 uV, v_off_fin = 2410.745 uV 20:25:09 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 173.236 uVs 20:25:11 I2Y Average and RMS of Measurements VT = -169.397 +- 6.768 uVs I2Y = -15.585 +- 0.623 uTm^2 20:25:19 Wire Position: X0 = 0.0000 mm, Y0 = -1.0000 mm 20:25:23 Integrated Voltage Measurement... 20:25:32 Voltage Offset Correction: v_off_ini = 2416.177 uV, v_off_fin = 2392.934 uV 20:25:32 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -108.730 uVs 20:25:32 Integrated Voltage Measurement... 20:25:41 Voltage Offset Correction: v_off_ini = 2393.048 uV, v_off_fin = 2368.361 uV 20:25:41 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 95.283 uVs 20:25:41 Integrated Voltage Measurement... 20:25:50 Voltage Offset Correction: v_off_ini = 2386.631 uV, v_off_fin = 2382.643 uV 20:25:50 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -120.343 uVs 20:25:50 Integrated Voltage Measurement... 20:25:59 Voltage Offset Correction: v_off_ini = 2397.753 uV, v_off_fin = 2389.033 uV 20:25:59 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 101.425 uVs 20:25:59 Integrated Voltage Measurement... 20:26:08 Voltage Offset Correction: v_off_ini = 2393.754 uV, v_off_fin = 2383.596 uV 20:26:08 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -95.379 uVs 20:26:08 Integrated Voltage Measurement... 20:26:17 Voltage Offset Correction: v_off_ini = 2405.608 uV, v_off_fin = 2392.128 uV 20:26:17 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 114.142 uVs 20:26:21 I1X Average and RMS of Measurements VT = -105.884 +- 3.710 uVs I1X = 21.177 +- 0.742 uTm 20:26:23 Wire Position: X0 = 0.0000 mm, Y0 = -1.0000 mm 20:26:28 Integrated Voltage Measurement... 20:26:37 Voltage Offset Correction: v_off_ini = 2397.776 uV, v_off_fin = 2379.251 uV 20:26:37 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -170.156 uVs 20:26:37 Integrated Voltage Measurement... 20:26:46 Voltage Offset Correction: v_off_ini = 2392.319 uV, v_off_fin = 2412.572 uV 20:26:46 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 161.425 uVs 20:26:46 Integrated Voltage Measurement... 20:26:55 Voltage Offset Correction: v_off_ini = 2414.620 uV, v_off_fin = 2390.745 uV 20:26:55 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -184.207 uVs 20:26:55 Integrated Voltage Measurement... 20:27:04 Voltage Offset Correction: v_off_ini = 2392.380 uV, v_off_fin = 2420.238 uV 20:27:04 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 190.585 uVs 20:27:04 Integrated Voltage Measurement... 20:27:12 Voltage Offset Correction: v_off_ini = 2412.094 uV, v_off_fin = 2409.256 uV 20:27:13 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -185.678 uVs 20:27:13 Integrated Voltage Measurement... 20:27:21 Voltage Offset Correction: v_off_ini = 2410.948 uV, v_off_fin = 2431.167 uV 20:27:21 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 170.245 uVs 20:27:26 I1Y Average and RMS of Measurements VT = -177.049 +- 8.844 uVs I1Y = -35.410 +- 1.769 uTm 20:27:28 Wire Position: X0 = 0.0000 mm, Y0 = -1.0000 mm 20:27:31 Integrated Voltage Measurement... 20:27:40 Voltage Offset Correction: v_off_ini = 2385.911 uV, v_off_fin = 2372.588 uV 20:27:40 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -79.930 uVs 20:27:40 Integrated Voltage Measurement... 20:27:49 Voltage Offset Correction: v_off_ini = 2427.520 uV, v_off_fin = 2419.907 uV 20:27:49 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 102.213 uVs 20:27:49 Integrated Voltage Measurement... 20:27:58 Voltage Offset Correction: v_off_ini = 2423.671 uV, v_off_fin = 2406.931 uV 20:27:58 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -104.378 uVs 20:27:58 Integrated Voltage Measurement... 20:28:07 Voltage Offset Correction: v_off_ini = 2434.579 uV, v_off_fin = 2442.814 uV 20:28:07 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 98.516 uVs 20:28:07 Integrated Voltage Measurement... 20:28:16 Voltage Offset Correction: v_off_ini = 2426.793 uV, v_off_fin = 2434.152 uV 20:28:16 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -123.234 uVs 20:28:16 Integrated Voltage Measurement... 20:28:25 Voltage Offset Correction: v_off_ini = 2441.277 uV, v_off_fin = 2416.257 uV 20:28:25 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 96.638 uVs 20:28:27 I2X Average and RMS of Measurements VT = -100.818 +- 7.714 uVs I2X = 9.275 +- 0.710 uTm^2 20:28:30 Wire Position: X0 = 0.0000 mm, Y0 = -1.0000 mm 20:28:32 Integrated Voltage Measurement... 20:28:41 Voltage Offset Correction: v_off_ini = 2429.144 uV, v_off_fin = 2433.145 uV 20:28:41 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -196.399 uVs 20:28:41 Integrated Voltage Measurement... 20:28:50 Voltage Offset Correction: v_off_ini = 2434.489 uV, v_off_fin = 2436.221 uV 20:28:50 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 179.698 uVs 20:28:50 Integrated Voltage Measurement... 20:28:59 Voltage Offset Correction: v_off_ini = 2442.322 uV, v_off_fin = 2434.096 uV 20:28:59 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -190.309 uVs 20:28:59 Integrated Voltage Measurement... 20:29:08 Voltage Offset Correction: v_off_ini = 2425.762 uV, v_off_fin = 2414.681 uV 20:29:08 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 181.811 uVs 20:29:08 Integrated Voltage Measurement... 20:29:17 Voltage Offset Correction: v_off_ini = 2433.757 uV, v_off_fin = 2439.262 uV 20:29:17 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -201.750 uVs 20:29:17 Integrated Voltage Measurement... 20:29:26 Voltage Offset Correction: v_off_ini = 2451.020 uV, v_off_fin = 2429.430 uV 20:29:26 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 193.935 uVs 20:29:29 I2Y Average and RMS of Measurements VT = -190.650 +- 5.150 uVs I2Y = -17.540 +- 0.474 uTm^2 20:29:33 Wire Position: X0 = 0.0000 mm, Y0 = -0.5000 mm 20:29:37 Integrated Voltage Measurement... 20:29:46 Voltage Offset Correction: v_off_ini = 2447.755 uV, v_off_fin = 2436.493 uV 20:29:46 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.4998 mm dXY = -0.4999 mm VT = -103.777 uVs 20:29:46 Integrated Voltage Measurement... 20:29:55 Voltage Offset Correction: v_off_ini = 2449.112 uV, v_off_fin = 2472.050 uV 20:29:55 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 106.167 uVs 20:29:55 Integrated Voltage Measurement... 20:30:04 Voltage Offset Correction: v_off_ini = 2479.226 uV, v_off_fin = 2466.456 uV 20:30:04 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -107.691 uVs 20:30:04 Integrated Voltage Measurement... 20:30:13 Voltage Offset Correction: v_off_ini = 2470.334 uV, v_off_fin = 2481.235 uV 20:30:13 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 101.332 uVs 20:30:13 Integrated Voltage Measurement... 20:30:22 Voltage Offset Correction: v_off_ini = 2471.572 uV, v_off_fin = 2480.849 uV 20:30:22 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -94.601 uVs 20:30:22 Integrated Voltage Measurement... 20:30:31 Voltage Offset Correction: v_off_ini = 2481.316 uV, v_off_fin = 2477.049 uV 20:30:31 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 77.551 uVs 20:30:35 I1X Average and RMS of Measurements VT = -98.520 +- 8.801 uVs I1X = 19.705 +- 1.761 uTm 20:30:38 Wire Position: X0 = 0.0000 mm, Y0 = -0.5000 mm 20:30:42 Integrated Voltage Measurement... 20:30:51 Voltage Offset Correction: v_off_ini = 2463.962 uV, v_off_fin = 2455.206 uV 20:30:51 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -164.015 uVs 20:30:51 Integrated Voltage Measurement... 20:31:00 Voltage Offset Correction: v_off_ini = 2479.176 uV, v_off_fin = 2476.806 uV 20:31:00 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 183.796 uVs 20:31:00 Integrated Voltage Measurement... 20:31:09 Voltage Offset Correction: v_off_ini = 2464.395 uV, v_off_fin = 2465.253 uV 20:31:09 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -174.300 uVs 20:31:09 Integrated Voltage Measurement... 20:31:18 Voltage Offset Correction: v_off_ini = 2460.364 uV, v_off_fin = 2482.608 uV 20:31:18 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 171.201 uVs 20:31:18 Integrated Voltage Measurement... 20:31:27 Voltage Offset Correction: v_off_ini = 2481.756 uV, v_off_fin = 2480.091 uV 20:31:27 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -177.864 uVs 20:31:27 Integrated Voltage Measurement... 20:31:36 Voltage Offset Correction: v_off_ini = 2480.640 uV, v_off_fin = 2482.014 uV 20:31:36 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 184.211 uVs 20:31:40 I1Y Average and RMS of Measurements VT = -175.898 +- 3.665 uVs I1Y = -35.180 +- 0.733 uTm 20:31:42 Wire Position: X0 = 0.0000 mm, Y0 = -0.5000 mm 20:31:45 Integrated Voltage Measurement... 20:31:54 Voltage Offset Correction: v_off_ini = 2460.154 uV, v_off_fin = 2445.014 uV 20:31:54 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -97.755 uVs 20:31:54 Integrated Voltage Measurement... 20:32:03 Voltage Offset Correction: v_off_ini = 2470.532 uV, v_off_fin = 2490.429 uV 20:32:03 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 95.413 uVs 20:32:03 Integrated Voltage Measurement... 20:32:12 Voltage Offset Correction: v_off_ini = 2460.567 uV, v_off_fin = 2459.824 uV 20:32:12 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -85.701 uVs 20:32:12 Integrated Voltage Measurement... 20:32:21 Voltage Offset Correction: v_off_ini = 2465.483 uV, v_off_fin = 2474.123 uV 20:32:21 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 81.402 uVs 20:32:21 Integrated Voltage Measurement... 20:32:30 Voltage Offset Correction: v_off_ini = 2458.353 uV, v_off_fin = 2466.229 uV 20:32:30 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -59.231 uVs 20:32:30 Integrated Voltage Measurement... 20:32:39 Voltage Offset Correction: v_off_ini = 2478.869 uV, v_off_fin = 2482.532 uV 20:32:39 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 102.248 uVs 20:32:41 I2X Average and RMS of Measurements VT = -86.958 +- 6.902 uVs I2X = 8.000 +- 0.635 uTm^2 20:32:44 Wire Position: X0 = 0.0000 mm, Y0 = -0.5000 mm 20:32:46 Integrated Voltage Measurement... 20:32:55 Voltage Offset Correction: v_off_ini = 2485.896 uV, v_off_fin = 2481.957 uV 20:32:55 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -173.856 uVs 20:32:55 Integrated Voltage Measurement... 20:33:04 Voltage Offset Correction: v_off_ini = 2495.159 uV, v_off_fin = 2485.403 uV 20:33:04 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 180.018 uVs 20:33:04 Integrated Voltage Measurement... 20:33:13 Voltage Offset Correction: v_off_ini = 2488.700 uV, v_off_fin = 2488.579 uV 20:33:13 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -166.468 uVs 20:33:13 Integrated Voltage Measurement... 20:33:22 Voltage Offset Correction: v_off_ini = 2489.424 uV, v_off_fin = 2496.562 uV 20:33:22 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 194.079 uVs 20:33:23 Integrated Voltage Measurement... 20:33:31 Voltage Offset Correction: v_off_ini = 2451.045 uV, v_off_fin = 2462.470 uV 20:33:31 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -198.938 uVs 20:33:32 Integrated Voltage Measurement... 20:33:40 Voltage Offset Correction: v_off_ini = 2454.931 uV, v_off_fin = 2466.983 uV 20:33:41 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 170.731 uVs 20:33:43 I2Y Average and RMS of Measurements VT = -180.682 +- 3.237 uVs I2Y = -16.623 +- 0.298 uTm^2 20:33:47 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 20:33:51 Integrated Voltage Measurement... 20:34:00 Voltage Offset Correction: v_off_ini = 2494.915 uV, v_off_fin = 2457.091 uV 20:34:00 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.4998 mm dXY = -0.4999 mm VT = -119.819 uVs 20:34:00 Integrated Voltage Measurement... 20:34:09 Voltage Offset Correction: v_off_ini = 2493.167 uV, v_off_fin = 2491.873 uV 20:34:09 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 93.833 uVs 20:34:09 Integrated Voltage Measurement... 20:34:18 Voltage Offset Correction: v_off_ini = 2485.600 uV, v_off_fin = 2474.507 uV 20:34:18 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -111.277 uVs 20:34:18 Integrated Voltage Measurement... 20:34:27 Voltage Offset Correction: v_off_ini = 2485.101 uV, v_off_fin = 2492.620 uV 20:34:27 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 102.336 uVs 20:34:27 Integrated Voltage Measurement... 20:34:36 Voltage Offset Correction: v_off_ini = 2480.809 uV, v_off_fin = 2468.388 uV 20:34:36 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -98.321 uVs 20:34:36 Integrated Voltage Measurement... 20:34:45 Voltage Offset Correction: v_off_ini = 2470.273 uV, v_off_fin = 2468.450 uV 20:34:45 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = 102.610 uVs 20:34:49 I1X Average and RMS of Measurements VT = -104.699 +- 2.994 uVs I1X = 20.941 +- 0.599 uTm 20:34:52 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 20:34:56 Integrated Voltage Measurement... 20:35:05 Voltage Offset Correction: v_off_ini = 2464.887 uV, v_off_fin = 2455.943 uV 20:35:05 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -155.448 uVs 20:35:05 Integrated Voltage Measurement... 20:35:14 Voltage Offset Correction: v_off_ini = 2473.851 uV, v_off_fin = 2498.368 uV 20:35:14 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 168.755 uVs 20:35:14 Integrated Voltage Measurement... 20:35:23 Voltage Offset Correction: v_off_ini = 2472.682 uV, v_off_fin = 2475.015 uV 20:35:23 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -164.528 uVs 20:35:23 Integrated Voltage Measurement... 20:35:32 Voltage Offset Correction: v_off_ini = 2483.510 uV, v_off_fin = 2490.795 uV 20:35:32 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 172.859 uVs 20:35:32 Integrated Voltage Measurement... 20:35:41 Voltage Offset Correction: v_off_ini = 2498.507 uV, v_off_fin = 2487.282 uV 20:35:41 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -176.907 uVs 20:35:41 Integrated Voltage Measurement... 20:35:50 Voltage Offset Correction: v_off_ini = 2491.929 uV, v_off_fin = 2484.402 uV 20:35:50 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 164.583 uVs 20:35:54 I1Y Average and RMS of Measurements VT = -167.180 +- 3.688 uVs I1Y = -33.436 +- 0.738 uTm 20:35:57 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 20:35:59 Integrated Voltage Measurement... 20:36:08 Voltage Offset Correction: v_off_ini = 2503.441 uV, v_off_fin = 2511.041 uV 20:36:08 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -72.916 uVs 20:36:08 Integrated Voltage Measurement... 20:36:17 Voltage Offset Correction: v_off_ini = 2512.025 uV, v_off_fin = 2524.657 uV 20:36:17 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 89.807 uVs 20:36:17 Integrated Voltage Measurement... 20:36:26 Voltage Offset Correction: v_off_ini = 2529.098 uV, v_off_fin = 2493.603 uV 20:36:26 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -69.257 uVs 20:36:26 Integrated Voltage Measurement... 20:36:35 Voltage Offset Correction: v_off_ini = 2522.796 uV, v_off_fin = 2528.135 uV 20:36:35 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 79.269 uVs 20:36:35 Integrated Voltage Measurement... 20:36:44 Voltage Offset Correction: v_off_ini = 2519.219 uV, v_off_fin = 2508.255 uV 20:36:44 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -89.061 uVs 20:36:44 Integrated Voltage Measurement... 20:36:53 Voltage Offset Correction: v_off_ini = 2498.467 uV, v_off_fin = 2475.702 uV 20:36:53 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 98.785 uVs 20:36:56 I2X Average and RMS of Measurements VT = -83.182 +- 8.129 uVs I2X = 7.653 +- 0.748 uTm^2 20:36:58 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 20:37:01 Integrated Voltage Measurement... 20:37:10 Voltage Offset Correction: v_off_ini = 2474.619 uV, v_off_fin = 2493.990 uV 20:37:10 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -152.515 uVs 20:37:10 Integrated Voltage Measurement... 20:37:19 Voltage Offset Correction: v_off_ini = 2478.376 uV, v_off_fin = 2485.583 uV 20:37:19 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 129.595 uVs 20:37:19 Integrated Voltage Measurement... 20:37:28 Voltage Offset Correction: v_off_ini = 2499.370 uV, v_off_fin = 2488.838 uV 20:37:28 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -167.076 uVs 20:37:28 Integrated Voltage Measurement... 20:37:37 Voltage Offset Correction: v_off_ini = 2496.614 uV, v_off_fin = 2473.830 uV 20:37:37 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 160.064 uVs 20:37:37 Integrated Voltage Measurement... 20:37:46 Voltage Offset Correction: v_off_ini = 2484.574 uV, v_off_fin = 2465.079 uV 20:37:46 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -184.451 uVs 20:37:46 Integrated Voltage Measurement... 20:37:55 Voltage Offset Correction: v_off_ini = 2496.404 uV, v_off_fin = 2496.900 uV 20:37:55 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 172.957 uVs 20:37:57 I2Y Average and RMS of Measurements VT = -161.110 +- 15.468 uVs I2Y = -14.822 +- 1.423 uTm^2 20:38:01 Wire Position: X0 = 0.0000 mm, Y0 = 0.5000 mm 20:38:05 Integrated Voltage Measurement... 20:38:14 Voltage Offset Correction: v_off_ini = 2469.405 uV, v_off_fin = 2448.620 uV 20:38:14 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -105.939 uVs 20:38:14 Integrated Voltage Measurement... 20:38:23 Voltage Offset Correction: v_off_ini = 2445.292 uV, v_off_fin = 2455.695 uV 20:38:23 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 89.994 uVs 20:38:23 Integrated Voltage Measurement... 20:38:32 Voltage Offset Correction: v_off_ini = 2448.204 uV, v_off_fin = 2454.760 uV 20:38:32 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -93.963 uVs 20:38:32 Integrated Voltage Measurement... 20:38:41 Voltage Offset Correction: v_off_ini = 2485.973 uV, v_off_fin = 2481.981 uV 20:38:41 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 106.291 uVs 20:38:41 Integrated Voltage Measurement... 20:38:50 Voltage Offset Correction: v_off_ini = 2456.693 uV, v_off_fin = 2463.674 uV 20:38:50 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -75.061 uVs 20:38:50 Integrated Voltage Measurement... 20:38:59 Voltage Offset Correction: v_off_ini = 2485.839 uV, v_off_fin = 2463.510 uV 20:38:59 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 115.918 uVs 20:39:04 I1X Average and RMS of Measurements VT = -97.861 +- 1.895 uVs I1X = 19.572 +- 0.379 uTm 20:39:06 Wire Position: X0 = 0.0000 mm, Y0 = 0.5000 mm 20:39:10 Integrated Voltage Measurement... 20:39:19 Voltage Offset Correction: v_off_ini = 2490.483 uV, v_off_fin = 2499.331 uV 20:39:19 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -136.585 uVs 20:39:19 Integrated Voltage Measurement... 20:39:28 Voltage Offset Correction: v_off_ini = 2502.650 uV, v_off_fin = 2520.546 uV 20:39:28 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 163.079 uVs 20:39:28 Integrated Voltage Measurement... 20:39:37 Voltage Offset Correction: v_off_ini = 2484.120 uV, v_off_fin = 2457.999 uV 20:39:37 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -158.396 uVs 20:39:37 Integrated Voltage Measurement... 20:39:46 Voltage Offset Correction: v_off_ini = 2451.369 uV, v_off_fin = 2472.900 uV 20:39:46 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 147.205 uVs 20:39:46 Integrated Voltage Measurement... 20:39:55 Voltage Offset Correction: v_off_ini = 2449.623 uV, v_off_fin = 2446.373 uV 20:39:55 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -143.557 uVs 20:39:55 Integrated Voltage Measurement... 20:40:04 Voltage Offset Correction: v_off_ini = 2471.367 uV, v_off_fin = 2468.683 uV 20:40:04 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 157.450 uVs 20:40:08 I1Y Average and RMS of Measurements VT = -151.045 +- 1.271 uVs I1Y = -30.208 +- 0.254 uTm 20:40:11 Wire Position: X0 = 0.0000 mm, Y0 = 0.5000 mm 20:40:13 Integrated Voltage Measurement... 20:40:22 Voltage Offset Correction: v_off_ini = 2447.865 uV, v_off_fin = 2446.449 uV 20:40:22 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -70.701 uVs 20:40:22 Integrated Voltage Measurement... 20:40:31 Voltage Offset Correction: v_off_ini = 2458.250 uV, v_off_fin = 2487.101 uV 20:40:31 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 69.710 uVs 20:40:31 Integrated Voltage Measurement... 20:40:40 Voltage Offset Correction: v_off_ini = 2463.408 uV, v_off_fin = 2461.766 uV 20:40:40 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -76.142 uVs 20:40:40 Integrated Voltage Measurement... 20:40:49 Voltage Offset Correction: v_off_ini = 2500.872 uV, v_off_fin = 2496.701 uV 20:40:49 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 72.895 uVs 20:40:50 Integrated Voltage Measurement... 20:40:58 Voltage Offset Correction: v_off_ini = 2472.876 uV, v_off_fin = 2459.007 uV 20:40:58 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -89.010 uVs 20:40:59 Integrated Voltage Measurement... 20:41:07 Voltage Offset Correction: v_off_ini = 2478.519 uV, v_off_fin = 2466.762 uV 20:41:08 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 67.724 uVs 20:41:10 I2X Average and RMS of Measurements VT = -74.364 +- 3.334 uVs I2X = 6.841 +- 0.307 uTm^2 20:41:12 Wire Position: X0 = 0.0000 mm, Y0 = 0.5000 mm 20:41:15 Integrated Voltage Measurement... 20:41:24 Voltage Offset Correction: v_off_ini = 2449.596 uV, v_off_fin = 2443.113 uV 20:41:24 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -159.033 uVs 20:41:24 Integrated Voltage Measurement... 20:41:33 Voltage Offset Correction: v_off_ini = 2439.065 uV, v_off_fin = 2453.889 uV 20:41:33 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 135.832 uVs 20:41:33 Integrated Voltage Measurement... 20:41:42 Voltage Offset Correction: v_off_ini = 2472.191 uV, v_off_fin = 2463.709 uV 20:41:42 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -157.966 uVs 20:41:42 Integrated Voltage Measurement... 20:41:51 Voltage Offset Correction: v_off_ini = 2459.823 uV, v_off_fin = 2478.822 uV 20:41:51 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 163.639 uVs 20:41:51 Integrated Voltage Measurement... 20:42:00 Voltage Offset Correction: v_off_ini = 2498.639 uV, v_off_fin = 2505.282 uV 20:42:00 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -149.304 uVs 20:42:00 Integrated Voltage Measurement... 20:42:09 Voltage Offset Correction: v_off_ini = 2501.810 uV, v_off_fin = 2511.759 uV 20:42:09 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 163.698 uVs 20:42:11 I2Y Average and RMS of Measurements VT = -154.912 +- 5.573 uVs I2Y = -14.252 +- 0.513 uTm^2 20:42:15 Wire Position: X0 = 0.0000 mm, Y0 = 1.0000 mm 20:42:20 Integrated Voltage Measurement... 20:42:29 Voltage Offset Correction: v_off_ini = 2524.001 uV, v_off_fin = 2496.485 uV 20:42:29 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -123.012 uVs 20:42:29 Integrated Voltage Measurement... 20:42:38 Voltage Offset Correction: v_off_ini = 2535.004 uV, v_off_fin = 2518.706 uV 20:42:38 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 76.427 uVs 20:42:38 Integrated Voltage Measurement... 20:42:47 Voltage Offset Correction: v_off_ini = 2508.092 uV, v_off_fin = 2500.109 uV 20:42:47 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -89.955 uVs 20:42:47 Integrated Voltage Measurement... 20:42:56 Voltage Offset Correction: v_off_ini = 2522.451 uV, v_off_fin = 2515.041 uV 20:42:56 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 81.127 uVs 20:42:56 Integrated Voltage Measurement... 20:43:05 Voltage Offset Correction: v_off_ini = 2505.163 uV, v_off_fin = 2501.925 uV 20:43:05 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -107.569 uVs 20:43:05 Integrated Voltage Measurement... 20:43:14 Voltage Offset Correction: v_off_ini = 2508.086 uV, v_off_fin = 2517.192 uV 20:43:14 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 90.565 uVs 20:43:18 I1X Average and RMS of Measurements VT = -94.776 +- 6.536 uVs I1X = 18.956 +- 1.307 uTm 20:43:20 Wire Position: X0 = 0.0000 mm, Y0 = 1.0000 mm 20:43:25 Integrated Voltage Measurement... 20:43:34 Voltage Offset Correction: v_off_ini = 2509.671 uV, v_off_fin = 2497.467 uV 20:43:34 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -139.738 uVs 20:43:34 Integrated Voltage Measurement... 20:43:43 Voltage Offset Correction: v_off_ini = 2546.714 uV, v_off_fin = 2557.700 uV 20:43:43 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 150.949 uVs 20:43:43 Integrated Voltage Measurement... 20:43:51 Voltage Offset Correction: v_off_ini = 2531.168 uV, v_off_fin = 2528.769 uV 20:43:51 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -150.111 uVs 20:43:52 Integrated Voltage Measurement... 20:44:00 Voltage Offset Correction: v_off_ini = 2533.633 uV, v_off_fin = 2543.967 uV 20:44:00 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 151.074 uVs 20:44:00 Integrated Voltage Measurement... 20:44:09 Voltage Offset Correction: v_off_ini = 2503.551 uV, v_off_fin = 2516.927 uV 20:44:09 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -129.133 uVs 20:44:09 Integrated Voltage Measurement... 20:44:18 Voltage Offset Correction: v_off_ini = 2547.117 uV, v_off_fin = 2548.031 uV 20:44:18 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 133.575 uVs 20:44:23 I1Y Average and RMS of Measurements VT = -142.430 +- 8.120 uVs I1Y = -28.485 +- 1.623 uTm 20:44:25 Wire Position: X0 = 0.0000 mm, Y0 = 1.0000 mm 20:44:28 Integrated Voltage Measurement... 20:44:37 Voltage Offset Correction: v_off_ini = 2539.298 uV, v_off_fin = 2529.893 uV 20:44:37 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -75.874 uVs 20:44:37 Integrated Voltage Measurement... 20:44:46 Voltage Offset Correction: v_off_ini = 2533.381 uV, v_off_fin = 2551.359 uV 20:44:46 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 39.753 uVs 20:44:46 Integrated Voltage Measurement... 20:44:55 Voltage Offset Correction: v_off_ini = 2548.973 uV, v_off_fin = 2514.935 uV 20:44:55 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -75.737 uVs 20:44:55 Integrated Voltage Measurement... 20:45:04 Voltage Offset Correction: v_off_ini = 2541.803 uV, v_off_fin = 2547.605 uV 20:45:04 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 75.195 uVs 20:45:04 Integrated Voltage Measurement... 20:45:13 Voltage Offset Correction: v_off_ini = 2542.300 uV, v_off_fin = 2555.235 uV 20:45:13 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -77.059 uVs 20:45:13 Integrated Voltage Measurement... 20:45:22 Voltage Offset Correction: v_off_ini = 2557.452 uV, v_off_fin = 2571.980 uV 20:45:22 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 89.463 uVs 20:45:24 I2X Average and RMS of Measurements VT = -72.180 +- 10.646 uVs I2X = 6.641 +- 0.979 uTm^2 20:45:27 Wire Position: X0 = 0.0000 mm, Y0 = 1.0000 mm 20:45:29 Integrated Voltage Measurement... 20:45:38 Voltage Offset Correction: v_off_ini = 2555.460 uV, v_off_fin = 2557.960 uV 20:45:38 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -125.219 uVs 20:45:38 Integrated Voltage Measurement... 20:45:47 Voltage Offset Correction: v_off_ini = 2548.594 uV, v_off_fin = 2554.013 uV 20:45:47 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 139.463 uVs 20:45:47 Integrated Voltage Measurement... 20:45:56 Voltage Offset Correction: v_off_ini = 2567.461 uV, v_off_fin = 2557.894 uV 20:45:56 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -155.168 uVs 20:45:56 Integrated Voltage Measurement... 20:46:05 Voltage Offset Correction: v_off_ini = 2550.830 uV, v_off_fin = 2553.275 uV 20:46:05 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 133.037 uVs 20:46:05 Integrated Voltage Measurement... 20:46:14 Voltage Offset Correction: v_off_ini = 2558.837 uV, v_off_fin = 2543.115 uV 20:46:14 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -135.829 uVs 20:46:14 Integrated Voltage Measurement... 20:46:23 Voltage Offset Correction: v_off_ini = 2543.638 uV, v_off_fin = 2538.142 uV 20:46:23 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 126.058 uVs 20:46:26 I2Y Average and RMS of Measurements VT = -135.796 +- 5.901 uVs I2Y = -12.493 +- 0.543 uTm^2 20:46:43 Wire Position: X0 = -2.0000 mm, Y0 = 0.0000 mm 20:46:47 Integrated Voltage Measurement... 20:46:56 Voltage Offset Correction: v_off_ini = 2549.436 uV, v_off_fin = 2547.743 uV 20:46:56 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -143.991 uVs 20:46:56 Integrated Voltage Measurement... 20:47:05 Voltage Offset Correction: v_off_ini = 2565.925 uV, v_off_fin = 2579.736 uV 20:47:05 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 141.594 uVs 20:47:05 Integrated Voltage Measurement... 20:47:14 Voltage Offset Correction: v_off_ini = 2574.926 uV, v_off_fin = 2566.065 uV 20:47:14 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -141.632 uVs 20:47:14 Integrated Voltage Measurement... 20:47:23 Voltage Offset Correction: v_off_ini = 2566.958 uV, v_off_fin = 2576.536 uV 20:47:23 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 122.079 uVs 20:47:23 Integrated Voltage Measurement... 20:47:32 Voltage Offset Correction: v_off_ini = 2574.169 uV, v_off_fin = 2565.030 uV 20:47:32 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -139.427 uVs 20:47:32 Integrated Voltage Measurement... 20:47:41 Voltage Offset Correction: v_off_ini = 2552.956 uV, v_off_fin = 2544.092 uV 20:47:41 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 126.966 uVs 20:47:45 I1X Average and RMS of Measurements VT = -135.948 +- 4.871 uVs I1X = 27.190 +- 0.974 uTm 20:47:48 Wire Position: X0 = -2.0000 mm, Y0 = 0.0000 mm 20:47:52 Integrated Voltage Measurement... 20:48:01 Voltage Offset Correction: v_off_ini = 2550.920 uV, v_off_fin = 2543.785 uV 20:48:01 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.4999 mm VT = -138.752 uVs 20:48:01 Integrated Voltage Measurement... 20:48:10 Voltage Offset Correction: v_off_ini = 2545.247 uV, v_off_fin = 2557.051 uV 20:48:10 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 148.443 uVs 20:48:10 Integrated Voltage Measurement... 20:48:19 Voltage Offset Correction: v_off_ini = 2548.453 uV, v_off_fin = 2534.039 uV 20:48:19 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -153.495 uVs 20:48:19 Integrated Voltage Measurement... 20:48:28 Voltage Offset Correction: v_off_ini = 2542.111 uV, v_off_fin = 2550.890 uV 20:48:28 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 137.577 uVs 20:48:28 Integrated Voltage Measurement... 20:48:37 Voltage Offset Correction: v_off_ini = 2531.317 uV, v_off_fin = 2523.234 uV 20:48:37 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -122.132 uVs 20:48:37 Integrated Voltage Measurement... 20:48:46 Voltage Offset Correction: v_off_ini = 2542.075 uV, v_off_fin = 2545.155 uV 20:48:46 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 134.228 uVs 20:48:50 I1Y Average and RMS of Measurements VT = -139.104 +- 7.765 uVs I1Y = -27.821 +- 1.553 uTm 20:48:55 Wire Position: X0 = -2.0000 mm, Y0 = 0.0000 mm 20:48:58 Integrated Voltage Measurement... 20:49:06 Voltage Offset Correction: v_off_ini = 2538.684 uV, v_off_fin = 2533.557 uV 20:49:07 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -151.786 uVs 20:49:07 Integrated Voltage Measurement... 20:49:16 Voltage Offset Correction: v_off_ini = 2574.719 uV, v_off_fin = 2578.259 uV 20:49:16 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 127.985 uVs 20:49:16 Integrated Voltage Measurement... 20:49:25 Voltage Offset Correction: v_off_ini = 2537.387 uV, v_off_fin = 2542.123 uV 20:49:25 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -143.283 uVs 20:49:25 Integrated Voltage Measurement... 20:49:34 Voltage Offset Correction: v_off_ini = 2598.262 uV, v_off_fin = 2614.572 uV 20:49:34 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 121.289 uVs 20:49:34 Integrated Voltage Measurement... 20:49:43 Voltage Offset Correction: v_off_ini = 2582.880 uV, v_off_fin = 2572.275 uV 20:49:43 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -126.104 uVs 20:49:43 Integrated Voltage Measurement... 20:49:52 Voltage Offset Correction: v_off_ini = 2547.946 uV, v_off_fin = 2574.296 uV 20:49:52 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 115.196 uVs 20:49:54 I2X Average and RMS of Measurements VT = -130.941 +- 7.910 uVs I2X = 12.047 +- 0.728 uTm^2 20:49:59 Wire Position: X0 = -2.0000 mm, Y0 = 0.0000 mm 20:50:02 Integrated Voltage Measurement... 20:50:11 Voltage Offset Correction: v_off_ini = 2552.826 uV, v_off_fin = 2562.175 uV 20:50:11 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -163.493 uVs 20:50:11 Integrated Voltage Measurement... 20:50:20 Voltage Offset Correction: v_off_ini = 2560.739 uV, v_off_fin = 2562.606 uV 20:50:20 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 163.851 uVs 20:50:20 Integrated Voltage Measurement... 20:50:29 Voltage Offset Correction: v_off_ini = 2545.934 uV, v_off_fin = 2559.697 uV 20:50:29 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -173.814 uVs 20:50:29 Integrated Voltage Measurement... 20:50:38 Voltage Offset Correction: v_off_ini = 2561.487 uV, v_off_fin = 2555.814 uV 20:50:38 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 111.848 uVs 20:50:38 Integrated Voltage Measurement... 20:50:47 Voltage Offset Correction: v_off_ini = 2555.749 uV, v_off_fin = 2565.350 uV 20:50:47 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -158.047 uVs 20:50:47 Integrated Voltage Measurement... 20:50:56 Voltage Offset Correction: v_off_ini = 2560.503 uV, v_off_fin = 2549.668 uV 20:50:56 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 135.405 uVs 20:50:58 I2Y Average and RMS of Measurements VT = -151.076 +- 9.047 uVs I2Y = -13.899 +- 0.832 uTm^2 20:51:03 Wire Position: X0 = -1.0000 mm, Y0 = 0.0000 mm 20:51:07 Integrated Voltage Measurement... 20:51:16 Voltage Offset Correction: v_off_ini = 2568.037 uV, v_off_fin = 2551.558 uV 20:51:16 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -117.367 uVs 20:51:16 Integrated Voltage Measurement... 20:51:25 Voltage Offset Correction: v_off_ini = 2561.210 uV, v_off_fin = 2569.063 uV 20:51:25 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = 137.746 uVs 20:51:25 Integrated Voltage Measurement... 20:51:34 Voltage Offset Correction: v_off_ini = 2605.601 uV, v_off_fin = 2578.405 uV 20:51:34 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -118.710 uVs 20:51:34 Integrated Voltage Measurement... 20:51:43 Voltage Offset Correction: v_off_ini = 2582.446 uV, v_off_fin = 2555.972 uV 20:51:43 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 124.598 uVs 20:51:43 Integrated Voltage Measurement... 20:51:52 Voltage Offset Correction: v_off_ini = 2585.335 uV, v_off_fin = 2573.343 uV 20:51:52 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -128.709 uVs 20:51:52 Integrated Voltage Measurement... 20:52:01 Voltage Offset Correction: v_off_ini = 2590.828 uV, v_off_fin = 2587.409 uV 20:52:01 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = 100.451 uVs 20:52:05 I1X Average and RMS of Measurements VT = -121.263 +- 5.305 uVs I1X = 24.254 +- 1.062 uTm 20:52:08 Wire Position: X0 = -1.0000 mm, Y0 = 0.0000 mm 20:52:12 Integrated Voltage Measurement... 20:52:21 Voltage Offset Correction: v_off_ini = 2578.442 uV, v_off_fin = 2575.726 uV 20:52:21 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -144.924 uVs 20:52:21 Integrated Voltage Measurement... 20:52:30 Voltage Offset Correction: v_off_ini = 2571.997 uV, v_off_fin = 2587.849 uV 20:52:30 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 142.342 uVs 20:52:30 Integrated Voltage Measurement... 20:52:39 Voltage Offset Correction: v_off_ini = 2559.060 uV, v_off_fin = 2553.029 uV 20:52:39 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -146.508 uVs 20:52:39 Integrated Voltage Measurement... 20:52:48 Voltage Offset Correction: v_off_ini = 2577.617 uV, v_off_fin = 2594.503 uV 20:52:48 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 139.386 uVs 20:52:48 Integrated Voltage Measurement... 20:52:57 Voltage Offset Correction: v_off_ini = 2587.323 uV, v_off_fin = 2578.724 uV 20:52:57 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -142.523 uVs 20:52:57 Integrated Voltage Measurement... 20:53:06 Voltage Offset Correction: v_off_ini = 2589.668 uV, v_off_fin = 2603.744 uV 20:53:06 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 151.917 uVs 20:53:10 I1Y Average and RMS of Measurements VT = -144.600 +- 1.874 uVs I1Y = -28.920 +- 0.374 uTm 20:53:15 Wire Position: X0 = -1.0000 mm, Y0 = 0.0000 mm 20:53:18 Integrated Voltage Measurement... 20:53:27 Voltage Offset Correction: v_off_ini = 2571.436 uV, v_off_fin = 2571.913 uV 20:53:27 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -84.481 uVs 20:53:27 Integrated Voltage Measurement... 20:53:36 Voltage Offset Correction: v_off_ini = 2594.460 uV, v_off_fin = 2588.403 uV 20:53:36 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 112.542 uVs 20:53:36 Integrated Voltage Measurement... 20:53:45 Voltage Offset Correction: v_off_ini = 2584.434 uV, v_off_fin = 2584.209 uV 20:53:45 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -109.542 uVs 20:53:45 Integrated Voltage Measurement... 20:53:54 Voltage Offset Correction: v_off_ini = 2583.125 uV, v_off_fin = 2600.478 uV 20:53:54 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 108.558 uVs 20:53:54 Integrated Voltage Measurement... 20:54:03 Voltage Offset Correction: v_off_ini = 2605.644 uV, v_off_fin = 2610.319 uV 20:54:03 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -92.409 uVs 20:54:03 Integrated Voltage Measurement... 20:54:12 Voltage Offset Correction: v_off_ini = 2622.220 uV, v_off_fin = 2621.579 uV 20:54:12 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 117.688 uVs 20:54:15 I2X Average and RMS of Measurements VT = -104.203 +- 4.344 uVs I2X = 9.587 +- 0.400 uTm^2 20:54:20 Wire Position: X0 = -1.0000 mm, Y0 = 0.0000 mm 20:54:22 Integrated Voltage Measurement... 20:54:31 Voltage Offset Correction: v_off_ini = 2605.486 uV, v_off_fin = 2600.710 uV 20:54:31 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -140.764 uVs 20:54:31 Integrated Voltage Measurement... 20:54:40 Voltage Offset Correction: v_off_ini = 2582.482 uV, v_off_fin = 2624.793 uV 20:54:40 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = 158.239 uVs 20:54:40 Integrated Voltage Measurement... 20:54:49 Voltage Offset Correction: v_off_ini = 2629.501 uV, v_off_fin = 2612.812 uV 20:54:49 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -136.130 uVs 20:54:49 Integrated Voltage Measurement... 20:54:58 Voltage Offset Correction: v_off_ini = 2598.932 uV, v_off_fin = 2596.397 uV 20:54:58 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 153.318 uVs 20:54:58 Integrated Voltage Measurement... 20:55:07 Voltage Offset Correction: v_off_ini = 2615.568 uV, v_off_fin = 2624.920 uV 20:55:07 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -153.821 uVs 20:55:07 Integrated Voltage Measurement... 20:55:16 Voltage Offset Correction: v_off_ini = 2625.660 uV, v_off_fin = 2610.212 uV 20:55:16 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 140.810 uVs 20:55:19 I2Y Average and RMS of Measurements VT = -147.181 +- 1.953 uVs I2Y = -13.541 +- 0.180 uTm^2 20:55:23 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 20:55:28 Integrated Voltage Measurement... 20:55:37 Voltage Offset Correction: v_off_ini = 2613.053 uV, v_off_fin = 2610.814 uV 20:55:37 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -103.721 uVs 20:55:37 Integrated Voltage Measurement... 20:55:46 Voltage Offset Correction: v_off_ini = 2631.062 uV, v_off_fin = 2625.712 uV 20:55:46 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 94.808 uVs 20:55:46 Integrated Voltage Measurement... 20:55:55 Voltage Offset Correction: v_off_ini = 2640.051 uV, v_off_fin = 2615.055 uV 20:55:55 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -113.166 uVs 20:55:55 Integrated Voltage Measurement... 20:56:04 Voltage Offset Correction: v_off_ini = 2615.351 uV, v_off_fin = 2626.083 uV 20:56:04 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 94.545 uVs 20:56:04 Integrated Voltage Measurement... 20:56:13 Voltage Offset Correction: v_off_ini = 2626.638 uV, v_off_fin = 2646.482 uV 20:56:13 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -103.287 uVs 20:56:13 Integrated Voltage Measurement... 20:56:22 Voltage Offset Correction: v_off_ini = 2651.946 uV, v_off_fin = 2631.793 uV 20:56:22 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 95.239 uVs 20:56:26 I1X Average and RMS of Measurements VT = -100.794 +- 2.165 uVs I1X = 20.159 +- 0.433 uTm 20:56:28 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 20:56:33 Integrated Voltage Measurement... 20:56:42 Voltage Offset Correction: v_off_ini = 2629.647 uV, v_off_fin = 2628.687 uV 20:56:42 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -153.136 uVs 20:56:42 Integrated Voltage Measurement... 20:56:51 Voltage Offset Correction: v_off_ini = 2640.417 uV, v_off_fin = 2676.594 uV 20:56:51 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 129.419 uVs 20:56:51 Integrated Voltage Measurement... 20:57:00 Voltage Offset Correction: v_off_ini = 2681.038 uV, v_off_fin = 2665.767 uV 20:57:00 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -145.767 uVs 20:57:00 Integrated Voltage Measurement... 20:57:09 Voltage Offset Correction: v_off_ini = 2658.453 uV, v_off_fin = 2656.246 uV 20:57:09 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 153.090 uVs 20:57:09 Integrated Voltage Measurement... 20:57:18 Voltage Offset Correction: v_off_ini = 2674.697 uV, v_off_fin = 2685.945 uV 20:57:18 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -140.476 uVs 20:57:18 Integrated Voltage Measurement... 20:57:26 Voltage Offset Correction: v_off_ini = 2681.080 uV, v_off_fin = 2681.449 uV 20:57:27 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 154.280 uVs 20:57:31 I1Y Average and RMS of Measurements VT = -146.028 +- 3.462 uVs I1Y = -29.206 +- 0.692 uTm 20:57:36 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 20:57:38 Integrated Voltage Measurement... 20:57:47 Voltage Offset Correction: v_off_ini = 2671.684 uV, v_off_fin = 2666.938 uV 20:57:47 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -77.079 uVs 20:57:47 Integrated Voltage Measurement... 20:57:56 Voltage Offset Correction: v_off_ini = 2657.214 uV, v_off_fin = 2689.248 uV 20:57:56 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 99.161 uVs 20:57:56 Integrated Voltage Measurement... 20:58:05 Voltage Offset Correction: v_off_ini = 2685.909 uV, v_off_fin = 2666.758 uV 20:58:05 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -77.494 uVs 20:58:05 Integrated Voltage Measurement... 20:58:14 Voltage Offset Correction: v_off_ini = 2661.177 uV, v_off_fin = 2671.802 uV 20:58:14 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 78.798 uVs 20:58:15 Integrated Voltage Measurement... 20:58:23 Voltage Offset Correction: v_off_ini = 2669.004 uV, v_off_fin = 2661.223 uV 20:58:24 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -71.764 uVs 20:58:24 Integrated Voltage Measurement... 20:58:33 Voltage Offset Correction: v_off_ini = 2636.838 uV, v_off_fin = 2664.187 uV 20:58:33 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 78.324 uVs 20:58:35 I2X Average and RMS of Measurements VT = -80.437 +- 5.579 uVs I2X = 7.400 +- 0.513 uTm^2 20:58:40 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 20:58:43 Integrated Voltage Measurement... 20:58:52 Voltage Offset Correction: v_off_ini = 2659.731 uV, v_off_fin = 2644.041 uV 20:58:52 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -139.553 uVs 20:58:52 Integrated Voltage Measurement... 20:59:01 Voltage Offset Correction: v_off_ini = 2644.204 uV, v_off_fin = 2644.372 uV 20:59:01 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 152.696 uVs 20:59:01 Integrated Voltage Measurement... 20:59:10 Voltage Offset Correction: v_off_ini = 2640.770 uV, v_off_fin = 2634.441 uV 20:59:10 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -164.201 uVs 20:59:10 Integrated Voltage Measurement... 20:59:19 Voltage Offset Correction: v_off_ini = 2642.209 uV, v_off_fin = 2642.305 uV 20:59:19 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 157.726 uVs 20:59:19 Integrated Voltage Measurement... 20:59:28 Voltage Offset Correction: v_off_ini = 2625.384 uV, v_off_fin = 2636.073 uV 20:59:28 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -147.003 uVs 20:59:28 Integrated Voltage Measurement... 20:59:37 Voltage Offset Correction: v_off_ini = 2652.290 uV, v_off_fin = 2658.159 uV 20:59:37 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 138.334 uVs 20:59:39 I2Y Average and RMS of Measurements VT = -149.919 +- 7.936 uVs I2Y = -13.793 +- 0.730 uTm^2 20:59:44 Wire Position: X0 = 1.0000 mm, Y0 = 0.0000 mm 20:59:48 Integrated Voltage Measurement... 20:59:57 Voltage Offset Correction: v_off_ini = 2644.322 uV, v_off_fin = 2659.042 uV 20:59:57 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -91.120 uVs 20:59:57 Integrated Voltage Measurement... 21:00:06 Voltage Offset Correction: v_off_ini = 2676.348 uV, v_off_fin = 2665.242 uV 21:00:06 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 57.730 uVs 21:00:06 Integrated Voltage Measurement... 21:00:15 Voltage Offset Correction: v_off_ini = 2661.021 uV, v_off_fin = 2653.778 uV 21:00:15 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -84.325 uVs 21:00:15 Integrated Voltage Measurement... 21:00:24 Voltage Offset Correction: v_off_ini = 2641.789 uV, v_off_fin = 2668.350 uV 21:00:24 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 85.742 uVs 21:00:24 Integrated Voltage Measurement... 21:00:33 Voltage Offset Correction: v_off_ini = 2652.981 uV, v_off_fin = 2654.051 uV 21:00:33 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -64.057 uVs 21:00:33 Integrated Voltage Measurement... 21:00:42 Voltage Offset Correction: v_off_ini = 2654.621 uV, v_off_fin = 2651.365 uV 21:00:42 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 71.708 uVs 21:00:46 I1X Average and RMS of Measurements VT = -75.780 +- 7.067 uVs I1X = 15.156 +- 1.413 uTm 21:00:49 Wire Position: X0 = 1.0000 mm, Y0 = 0.0000 mm 21:00:53 Integrated Voltage Measurement... 21:01:02 Voltage Offset Correction: v_off_ini = 2652.988 uV, v_off_fin = 2640.395 uV 21:01:02 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -156.734 uVs 21:01:02 Integrated Voltage Measurement... 21:01:11 Voltage Offset Correction: v_off_ini = 2663.421 uV, v_off_fin = 2688.049 uV 21:01:11 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 161.825 uVs 21:01:11 Integrated Voltage Measurement... 21:01:20 Voltage Offset Correction: v_off_ini = 2685.814 uV, v_off_fin = 2671.033 uV 21:01:20 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -150.127 uVs 21:01:20 Integrated Voltage Measurement... 21:01:29 Voltage Offset Correction: v_off_ini = 2665.430 uV, v_off_fin = 2658.785 uV 21:01:29 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 158.384 uVs 21:01:29 Integrated Voltage Measurement... 21:01:38 Voltage Offset Correction: v_off_ini = 2660.529 uV, v_off_fin = 2669.065 uV 21:01:38 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -159.703 uVs 21:01:38 Integrated Voltage Measurement... 21:01:47 Voltage Offset Correction: v_off_ini = 2643.111 uV, v_off_fin = 2643.786 uV 21:01:47 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 150.838 uVs 21:01:51 I1Y Average and RMS of Measurements VT = -156.269 +- 2.169 uVs I1Y = -31.254 +- 0.434 uTm 21:01:56 Wire Position: X0 = 1.0000 mm, Y0 = 0.0000 mm 21:01:59 Integrated Voltage Measurement... 21:02:08 Voltage Offset Correction: v_off_ini = 2639.070 uV, v_off_fin = 2612.733 uV 21:02:08 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -56.452 uVs 21:02:08 Integrated Voltage Measurement... 21:02:17 Voltage Offset Correction: v_off_ini = 2630.858 uV, v_off_fin = 2643.395 uV 21:02:17 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 81.772 uVs 21:02:17 Integrated Voltage Measurement... 21:02:26 Voltage Offset Correction: v_off_ini = 2647.896 uV, v_off_fin = 2632.614 uV 21:02:26 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -70.048 uVs 21:02:26 Integrated Voltage Measurement... 21:02:35 Voltage Offset Correction: v_off_ini = 2633.575 uV, v_off_fin = 2642.315 uV 21:02:35 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 52.092 uVs 21:02:35 Integrated Voltage Measurement... 21:02:44 Voltage Offset Correction: v_off_ini = 2633.928 uV, v_off_fin = 2633.574 uV 21:02:44 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -70.939 uVs 21:02:44 Integrated Voltage Measurement... 21:02:53 Voltage Offset Correction: v_off_ini = 2620.370 uV, v_off_fin = 2628.205 uV 21:02:53 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 67.691 uVs 21:02:55 I2X Average and RMS of Measurements VT = -66.499 +- 3.840 uVs I2X = 6.118 +- 0.353 uTm^2 21:03:00 Wire Position: X0 = 1.0000 mm, Y0 = 0.0000 mm 21:03:03 Integrated Voltage Measurement... 21:03:12 Voltage Offset Correction: v_off_ini = 2621.331 uV, v_off_fin = 2619.386 uV 21:03:12 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -151.200 uVs 21:03:12 Integrated Voltage Measurement... 21:03:21 Voltage Offset Correction: v_off_ini = 2629.395 uV, v_off_fin = 2621.076 uV 21:03:21 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 172.684 uVs 21:03:21 Integrated Voltage Measurement... 21:03:30 Voltage Offset Correction: v_off_ini = 2646.455 uV, v_off_fin = 2632.221 uV 21:03:30 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -161.028 uVs 21:03:30 Integrated Voltage Measurement... 21:03:39 Voltage Offset Correction: v_off_ini = 2630.301 uV, v_off_fin = 2628.622 uV 21:03:39 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 163.295 uVs 21:03:39 Integrated Voltage Measurement... 21:03:48 Voltage Offset Correction: v_off_ini = 2646.248 uV, v_off_fin = 2647.620 uV 21:03:48 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -153.749 uVs 21:03:48 Integrated Voltage Measurement... 21:03:57 Voltage Offset Correction: v_off_ini = 2632.224 uV, v_off_fin = 2631.722 uV 21:03:57 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 161.059 uVs 21:03:59 I2Y Average and RMS of Measurements VT = -160.503 +- 2.193 uVs I2Y = -14.766 +- 0.202 uTm^2 21:04:04 Wire Position: X0 = 2.0000 mm, Y0 = 0.0000 mm 21:04:08 Integrated Voltage Measurement... 21:04:17 Voltage Offset Correction: v_off_ini = 2645.028 uV, v_off_fin = 2651.221 uV 21:04:17 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -83.753 uVs 21:04:17 Integrated Voltage Measurement... 21:04:26 Voltage Offset Correction: v_off_ini = 2664.384 uV, v_off_fin = 2660.718 uV 21:04:26 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 50.161 uVs 21:04:26 Integrated Voltage Measurement... 21:04:35 Voltage Offset Correction: v_off_ini = 2631.538 uV, v_off_fin = 2671.455 uV 21:04:35 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -47.200 uVs 21:04:35 Integrated Voltage Measurement... 21:04:44 Voltage Offset Correction: v_off_ini = 2671.132 uV, v_off_fin = 2674.000 uV 21:04:44 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 68.618 uVs 21:04:44 Integrated Voltage Measurement... 21:04:53 Voltage Offset Correction: v_off_ini = 2646.824 uV, v_off_fin = 2648.597 uV 21:04:53 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -73.921 uVs 21:04:53 Integrated Voltage Measurement... 21:05:02 Voltage Offset Correction: v_off_ini = 2635.051 uV, v_off_fin = 2629.160 uV 21:05:02 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 67.229 uVs 21:05:07 I1X Average and RMS of Measurements VT = -65.147 +- 5.327 uVs I1X = 13.030 +- 1.066 uTm 21:05:09 Wire Position: X0 = 2.0000 mm, Y0 = 0.0000 mm 21:05:13 Integrated Voltage Measurement... 21:05:22 Voltage Offset Correction: v_off_ini = 2658.439 uV, v_off_fin = 2648.249 uV 21:05:22 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -187.682 uVs 21:05:22 Integrated Voltage Measurement... 21:05:31 Voltage Offset Correction: v_off_ini = 2661.617 uV, v_off_fin = 2673.181 uV 21:05:31 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 155.421 uVs 21:05:31 Integrated Voltage Measurement... 21:05:40 Voltage Offset Correction: v_off_ini = 2622.755 uV, v_off_fin = 2608.519 uV 21:05:40 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -153.583 uVs 21:05:40 Integrated Voltage Measurement... 21:05:49 Voltage Offset Correction: v_off_ini = 2616.570 uV, v_off_fin = 2630.685 uV 21:05:49 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 170.915 uVs 21:05:49 Integrated Voltage Measurement... 21:05:58 Voltage Offset Correction: v_off_ini = 2640.229 uV, v_off_fin = 2622.696 uV 21:05:58 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -171.422 uVs 21:05:58 Integrated Voltage Measurement... 21:06:07 Voltage Offset Correction: v_off_ini = 2616.629 uV, v_off_fin = 2635.428 uV 21:06:07 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 186.287 uVs 21:06:11 I1Y Average and RMS of Measurements VT = -170.885 +- 6.795 uVs I1Y = -34.177 +- 1.359 uTm 21:06:16 Wire Position: X0 = 2.0000 mm, Y0 = 0.0000 mm 21:06:19 Integrated Voltage Measurement... 21:06:28 Voltage Offset Correction: v_off_ini = 2624.422 uV, v_off_fin = 2626.592 uV 21:06:28 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -42.806 uVs 21:06:28 Integrated Voltage Measurement... 21:06:37 Voltage Offset Correction: v_off_ini = 2637.044 uV, v_off_fin = 2647.023 uV 21:06:37 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 55.690 uVs 21:06:37 Integrated Voltage Measurement... 21:06:46 Voltage Offset Correction: v_off_ini = 2625.079 uV, v_off_fin = 2630.190 uV 21:06:46 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -30.721 uVs 21:06:46 Integrated Voltage Measurement... 21:06:55 Voltage Offset Correction: v_off_ini = 2627.906 uV, v_off_fin = 2637.118 uV 21:06:55 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 45.493 uVs 21:06:55 Integrated Voltage Measurement... 21:07:04 Voltage Offset Correction: v_off_ini = 2643.146 uV, v_off_fin = 2647.429 uV 21:07:04 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -21.013 uVs 21:07:04 Integrated Voltage Measurement... 21:07:13 Voltage Offset Correction: v_off_ini = 2658.791 uV, v_off_fin = 2652.446 uV 21:07:13 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 25.616 uVs 21:07:16 I2X Average and RMS of Measurements VT = -36.890 +- 10.622 uVs I2X = 3.394 +- 0.977 uTm^2 21:07:21 Wire Position: X0 = 2.0000 mm, Y0 = 0.0000 mm 21:07:23 Integrated Voltage Measurement... 21:07:32 Voltage Offset Correction: v_off_ini = 2656.921 uV, v_off_fin = 2657.385 uV 21:07:32 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -175.046 uVs 21:07:32 Integrated Voltage Measurement... 21:07:41 Voltage Offset Correction: v_off_ini = 2656.349 uV, v_off_fin = 2679.366 uV 21:07:41 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 170.405 uVs 21:07:41 Integrated Voltage Measurement... 21:07:50 Voltage Offset Correction: v_off_ini = 2649.372 uV, v_off_fin = 2673.942 uV 21:07:50 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -166.697 uVs 21:07:50 Integrated Voltage Measurement... 21:07:59 Voltage Offset Correction: v_off_ini = 2680.612 uV, v_off_fin = 2674.547 uV 21:07:59 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 163.012 uVs 21:08:00 Integrated Voltage Measurement... 21:08:08 Voltage Offset Correction: v_off_ini = 2687.667 uV, v_off_fin = 2694.954 uV 21:08:09 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -174.956 uVs 21:08:09 Integrated Voltage Measurement... 21:08:18 Voltage Offset Correction: v_off_ini = 2679.632 uV, v_off_fin = 2676.041 uV 21:08:18 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 168.548 uVs 21:08:20 I2Y Average and RMS of Measurements VT = -169.777 +- 3.504 uVs I2Y = -15.620 +- 0.322 uTm^2 21:08:28 Wire Position: X0 = 0.0000 mm, Y0 = -1.0000 mm 21:08:32 Integrated Voltage Measurement... 21:08:41 Voltage Offset Correction: v_off_ini = 2680.637 uV, v_off_fin = 2667.147 uV 21:08:41 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -88.765 uVs 21:08:41 Integrated Voltage Measurement... 21:08:50 Voltage Offset Correction: v_off_ini = 2663.798 uV, v_off_fin = 2647.353 uV 21:08:50 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 108.308 uVs 21:08:50 Integrated Voltage Measurement... 21:08:59 Voltage Offset Correction: v_off_ini = 2662.116 uV, v_off_fin = 2649.917 uV 21:08:59 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -103.151 uVs 21:08:59 Integrated Voltage Measurement... 21:09:08 Voltage Offset Correction: v_off_ini = 2668.325 uV, v_off_fin = 2662.560 uV 21:09:08 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 112.914 uVs 21:09:08 Integrated Voltage Measurement... 21:09:17 Voltage Offset Correction: v_off_ini = 2663.835 uV, v_off_fin = 2655.320 uV 21:09:17 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -118.080 uVs 21:09:17 Integrated Voltage Measurement... 21:09:26 Voltage Offset Correction: v_off_ini = 2691.797 uV, v_off_fin = 2664.791 uV 21:09:26 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 130.497 uVs 21:09:30 I1X Average and RMS of Measurements VT = -110.286 +- 10.633 uVs I1X = 22.057 +- 2.126 uTm 21:09:33 Wire Position: X0 = 0.0000 mm, Y0 = -1.0000 mm 21:09:37 Integrated Voltage Measurement... 21:09:46 Voltage Offset Correction: v_off_ini = 2660.913 uV, v_off_fin = 2638.049 uV 21:09:46 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -187.655 uVs 21:09:46 Integrated Voltage Measurement... 21:09:55 Voltage Offset Correction: v_off_ini = 2645.871 uV, v_off_fin = 2662.503 uV 21:09:55 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 182.322 uVs 21:09:55 Integrated Voltage Measurement... 21:10:04 Voltage Offset Correction: v_off_ini = 2644.408 uV, v_off_fin = 2653.582 uV 21:10:04 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -157.198 uVs 21:10:04 Integrated Voltage Measurement... 21:10:13 Voltage Offset Correction: v_off_ini = 2662.644 uV, v_off_fin = 2659.034 uV 21:10:13 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 186.850 uVs 21:10:13 Integrated Voltage Measurement... 21:10:22 Voltage Offset Correction: v_off_ini = 2664.954 uV, v_off_fin = 2665.258 uV 21:10:22 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -174.606 uVs 21:10:22 Integrated Voltage Measurement... 21:10:31 Voltage Offset Correction: v_off_ini = 2669.892 uV, v_off_fin = 2669.195 uV 21:10:31 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 170.121 uVs 21:10:35 I1Y Average and RMS of Measurements VT = -176.459 +- 6.033 uVs I1Y = -35.292 +- 1.207 uTm 21:10:37 Wire Position: X0 = 0.0000 mm, Y0 = -1.0000 mm 21:10:40 Integrated Voltage Measurement... 21:10:49 Voltage Offset Correction: v_off_ini = 2646.546 uV, v_off_fin = 2656.736 uV 21:10:49 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -77.615 uVs 21:10:49 Integrated Voltage Measurement... 21:10:58 Voltage Offset Correction: v_off_ini = 2668.966 uV, v_off_fin = 2672.120 uV 21:10:58 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 81.269 uVs 21:10:58 Integrated Voltage Measurement... 21:11:07 Voltage Offset Correction: v_off_ini = 2676.352 uV, v_off_fin = 2632.483 uV 21:11:07 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -93.567 uVs 21:11:07 Integrated Voltage Measurement... 21:11:16 Voltage Offset Correction: v_off_ini = 2652.312 uV, v_off_fin = 2661.033 uV 21:11:16 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 91.351 uVs 21:11:16 Integrated Voltage Measurement... 21:11:25 Voltage Offset Correction: v_off_ini = 2659.447 uV, v_off_fin = 2673.800 uV 21:11:25 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -108.951 uVs 21:11:25 Integrated Voltage Measurement... 21:11:34 Voltage Offset Correction: v_off_ini = 2693.804 uV, v_off_fin = 2686.872 uV 21:11:34 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 83.044 uVs 21:11:37 I2X Average and RMS of Measurements VT = -89.299 +- 7.118 uVs I2X = 8.216 +- 0.655 uTm^2 21:11:39 Wire Position: X0 = 0.0000 mm, Y0 = -1.0000 mm 21:11:42 Integrated Voltage Measurement... 21:11:51 Voltage Offset Correction: v_off_ini = 2662.404 uV, v_off_fin = 2663.906 uV 21:11:51 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -166.821 uVs 21:11:51 Integrated Voltage Measurement... 21:12:00 Voltage Offset Correction: v_off_ini = 2658.237 uV, v_off_fin = 2674.225 uV 21:12:00 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 195.393 uVs 21:12:00 Integrated Voltage Measurement... 21:12:09 Voltage Offset Correction: v_off_ini = 2674.513 uV, v_off_fin = 2653.933 uV 21:12:09 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -162.418 uVs 21:12:09 Integrated Voltage Measurement... 21:12:18 Voltage Offset Correction: v_off_ini = 2668.337 uV, v_off_fin = 2675.771 uV 21:12:18 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 159.146 uVs 21:12:18 Integrated Voltage Measurement... 21:12:27 Voltage Offset Correction: v_off_ini = 2695.117 uV, v_off_fin = 2690.133 uV 21:12:27 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -175.123 uVs 21:12:27 Integrated Voltage Measurement... 21:12:36 Voltage Offset Correction: v_off_ini = 2691.573 uV, v_off_fin = 2691.120 uV 21:12:36 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 179.144 uVs 21:12:38 I2Y Average and RMS of Measurements VT = -173.008 +- 8.796 uVs I2Y = -15.917 +- 0.809 uTm^2 21:12:42 Wire Position: X0 = 0.0000 mm, Y0 = -0.5000 mm 21:12:47 Integrated Voltage Measurement... 21:12:56 Voltage Offset Correction: v_off_ini = 2670.996 uV, v_off_fin = 2680.189 uV 21:12:56 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -99.071 uVs 21:12:56 Integrated Voltage Measurement... 21:13:05 Voltage Offset Correction: v_off_ini = 2671.968 uV, v_off_fin = 2673.443 uV 21:13:05 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = 107.470 uVs 21:13:05 Integrated Voltage Measurement... 21:13:14 Voltage Offset Correction: v_off_ini = 2666.475 uV, v_off_fin = 2694.641 uV 21:13:14 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -96.728 uVs 21:13:14 Integrated Voltage Measurement... 21:13:23 Voltage Offset Correction: v_off_ini = 2700.638 uV, v_off_fin = 2692.663 uV 21:13:23 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 66.498 uVs 21:13:23 Integrated Voltage Measurement... 21:13:32 Voltage Offset Correction: v_off_ini = 2681.539 uV, v_off_fin = 2667.878 uV 21:13:32 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -86.465 uVs 21:13:32 Integrated Voltage Measurement... 21:13:41 Voltage Offset Correction: v_off_ini = 2668.685 uV, v_off_fin = 2673.346 uV 21:13:41 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 93.262 uVs 21:13:45 I1X Average and RMS of Measurements VT = -91.582 +- 8.925 uVs I1X = 18.317 +- 1.786 uTm 21:13:47 Wire Position: X0 = 0.0000 mm, Y0 = -0.5000 mm 21:13:52 Integrated Voltage Measurement... 21:14:00 Voltage Offset Correction: v_off_ini = 2656.470 uV, v_off_fin = 2648.772 uV 21:14:00 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -175.907 uVs 21:14:01 Integrated Voltage Measurement... 21:14:09 Voltage Offset Correction: v_off_ini = 2659.690 uV, v_off_fin = 2673.258 uV 21:14:09 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 164.444 uVs 21:14:10 Integrated Voltage Measurement... 21:14:18 Voltage Offset Correction: v_off_ini = 2668.216 uV, v_off_fin = 2650.506 uV 21:14:18 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -157.437 uVs 21:14:19 Integrated Voltage Measurement... 21:14:27 Voltage Offset Correction: v_off_ini = 2661.053 uV, v_off_fin = 2668.204 uV 21:14:27 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 151.654 uVs 21:14:28 Integrated Voltage Measurement... 21:14:36 Voltage Offset Correction: v_off_ini = 2666.830 uV, v_off_fin = 2649.014 uV 21:14:36 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -168.159 uVs 21:14:36 Integrated Voltage Measurement... 21:14:45 Voltage Offset Correction: v_off_ini = 2663.049 uV, v_off_fin = 2666.134 uV 21:14:45 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 184.778 uVs 21:14:50 I1Y Average and RMS of Measurements VT = -167.063 +- 9.217 uVs I1Y = -33.413 +- 1.843 uTm 21:14:52 Wire Position: X0 = 0.0000 mm, Y0 = -0.5000 mm 21:14:55 Integrated Voltage Measurement... 21:15:04 Voltage Offset Correction: v_off_ini = 2626.956 uV, v_off_fin = 2636.730 uV 21:15:04 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -55.163 uVs 21:15:04 Integrated Voltage Measurement... 21:15:13 Voltage Offset Correction: v_off_ini = 2687.048 uV, v_off_fin = 2707.683 uV 21:15:13 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 76.447 uVs 21:15:13 Integrated Voltage Measurement... 21:15:22 Voltage Offset Correction: v_off_ini = 2667.116 uV, v_off_fin = 2669.266 uV 21:15:22 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -70.474 uVs 21:15:22 Integrated Voltage Measurement... 21:15:31 Voltage Offset Correction: v_off_ini = 2681.157 uV, v_off_fin = 2683.903 uV 21:15:31 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 97.755 uVs 21:15:31 Integrated Voltage Measurement... 21:15:40 Voltage Offset Correction: v_off_ini = 2680.539 uV, v_off_fin = 2670.000 uV 21:15:40 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -74.844 uVs 21:15:40 Integrated Voltage Measurement... 21:15:49 Voltage Offset Correction: v_off_ini = 2690.056 uV, v_off_fin = 2709.534 uV 21:15:49 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 91.959 uVs 21:15:51 I2X Average and RMS of Measurements VT = -77.774 +- 8.468 uVs I2X = 7.155 +- 0.779 uTm^2 21:15:54 Wire Position: X0 = 0.0000 mm, Y0 = -0.5000 mm 21:15:56 Integrated Voltage Measurement... 21:16:05 Voltage Offset Correction: v_off_ini = 2712.645 uV, v_off_fin = 2708.285 uV 21:16:05 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -166.720 uVs 21:16:05 Integrated Voltage Measurement... 21:16:14 Voltage Offset Correction: v_off_ini = 2699.341 uV, v_off_fin = 2698.453 uV 21:16:14 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 154.353 uVs 21:16:14 Integrated Voltage Measurement... 21:16:23 Voltage Offset Correction: v_off_ini = 2694.332 uV, v_off_fin = 2697.261 uV 21:16:23 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -157.604 uVs 21:16:23 Integrated Voltage Measurement... 21:16:32 Voltage Offset Correction: v_off_ini = 2700.428 uV, v_off_fin = 2688.970 uV 21:16:32 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 168.533 uVs 21:16:33 Integrated Voltage Measurement... 21:16:41 Voltage Offset Correction: v_off_ini = 2684.043 uV, v_off_fin = 2687.396 uV 21:16:42 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -157.380 uVs 21:16:42 Integrated Voltage Measurement... 21:16:51 Voltage Offset Correction: v_off_ini = 2709.802 uV, v_off_fin = 2731.341 uV 21:16:51 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 183.260 uVs 21:16:53 I2Y Average and RMS of Measurements VT = -164.642 +- 4.146 uVs I2Y = -15.147 +- 0.381 uTm^2 21:16:57 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 21:17:01 Integrated Voltage Measurement... 21:17:10 Voltage Offset Correction: v_off_ini = 2709.634 uV, v_off_fin = 2690.276 uV 21:17:10 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.4999 mm dXY = -0.4999 mm VT = -127.228 uVs 21:17:10 Integrated Voltage Measurement... 21:17:19 Voltage Offset Correction: v_off_ini = 2710.148 uV, v_off_fin = 2726.474 uV 21:17:19 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = 98.943 uVs 21:17:19 Integrated Voltage Measurement... 21:17:28 Voltage Offset Correction: v_off_ini = 2737.081 uV, v_off_fin = 2724.415 uV 21:17:28 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -112.376 uVs 21:17:28 Integrated Voltage Measurement... 21:17:37 Voltage Offset Correction: v_off_ini = 2722.339 uV, v_off_fin = 2728.670 uV 21:17:37 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 77.780 uVs 21:17:37 Integrated Voltage Measurement... 21:17:46 Voltage Offset Correction: v_off_ini = 2713.660 uV, v_off_fin = 2724.865 uV 21:17:46 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -96.063 uVs 21:17:46 Integrated Voltage Measurement... 21:17:55 Voltage Offset Correction: v_off_ini = 2719.251 uV, v_off_fin = 2722.553 uV 21:17:55 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 108.256 uVs 21:18:00 I1X Average and RMS of Measurements VT = -103.441 +- 7.407 uVs I1X = 20.689 +- 1.483 uTm 21:18:02 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 21:18:06 Integrated Voltage Measurement... 21:18:15 Voltage Offset Correction: v_off_ini = 2733.725 uV, v_off_fin = 2732.069 uV 21:18:15 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -141.963 uVs 21:18:15 Integrated Voltage Measurement... 21:18:24 Voltage Offset Correction: v_off_ini = 2754.565 uV, v_off_fin = 2762.261 uV 21:18:24 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 170.598 uVs 21:18:24 Integrated Voltage Measurement... 21:18:33 Voltage Offset Correction: v_off_ini = 2754.290 uV, v_off_fin = 2739.805 uV 21:18:33 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -159.106 uVs 21:18:33 Integrated Voltage Measurement... 21:18:42 Voltage Offset Correction: v_off_ini = 2734.885 uV, v_off_fin = 2731.885 uV 21:18:42 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 122.026 uVs 21:18:42 Integrated Voltage Measurement... 21:18:51 Voltage Offset Correction: v_off_ini = 2754.171 uV, v_off_fin = 2725.991 uV 21:18:51 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -158.834 uVs 21:18:51 Integrated Voltage Measurement... 21:19:00 Voltage Offset Correction: v_off_ini = 2772.129 uV, v_off_fin = 2758.777 uV 21:19:00 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 165.258 uVs 21:19:04 I1Y Average and RMS of Measurements VT = -152.964 +- 9.077 uVs I1Y = -30.593 +- 1.815 uTm 21:19:07 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 21:19:09 Integrated Voltage Measurement... 21:19:18 Voltage Offset Correction: v_off_ini = 2779.049 uV, v_off_fin = 2747.658 uV 21:19:19 Integrated Voltage Measurement dXYus = -1.0001 mm, dXYds = 0.0000 mm dXY = -1.0001 mm VT = -75.803 uVs 21:19:19 Integrated Voltage Measurement... 21:19:27 Voltage Offset Correction: v_off_ini = 2761.925 uV, v_off_fin = 2772.475 uV 21:19:28 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 96.322 uVs 21:19:28 Integrated Voltage Measurement... 21:19:37 Voltage Offset Correction: v_off_ini = 2769.299 uV, v_off_fin = 2748.446 uV 21:19:37 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -74.870 uVs 21:19:37 Integrated Voltage Measurement... 21:19:46 Voltage Offset Correction: v_off_ini = 2782.184 uV, v_off_fin = 2786.442 uV 21:19:46 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 75.925 uVs 21:19:46 Integrated Voltage Measurement... 21:19:55 Voltage Offset Correction: v_off_ini = 2777.788 uV, v_off_fin = 2781.078 uV 21:19:55 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -87.357 uVs 21:19:55 Integrated Voltage Measurement... 21:20:04 Voltage Offset Correction: v_off_ini = 2758.248 uV, v_off_fin = 2758.670 uV 21:20:04 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 90.280 uVs 21:20:06 I2X Average and RMS of Measurements VT = -83.426 +- 5.787 uVs I2X = 7.675 +- 0.532 uTm^2 21:20:09 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 21:20:11 Integrated Voltage Measurement... 21:20:20 Voltage Offset Correction: v_off_ini = 2773.902 uV, v_off_fin = 2803.908 uV 21:20:20 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -160.273 uVs 21:20:20 Integrated Voltage Measurement... 21:20:29 Voltage Offset Correction: v_off_ini = 2791.411 uV, v_off_fin = 2768.799 uV 21:20:29 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 154.538 uVs 21:20:29 Integrated Voltage Measurement... 21:20:38 Voltage Offset Correction: v_off_ini = 2789.546 uV, v_off_fin = 2783.500 uV 21:20:38 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -149.961 uVs 21:20:38 Integrated Voltage Measurement... 21:20:47 Voltage Offset Correction: v_off_ini = 2796.632 uV, v_off_fin = 2790.840 uV 21:20:47 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 149.547 uVs 21:20:47 Integrated Voltage Measurement... 21:20:56 Voltage Offset Correction: v_off_ini = 2778.637 uV, v_off_fin = 2771.083 uV 21:20:56 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -175.360 uVs 21:20:56 Integrated Voltage Measurement... 21:21:05 Voltage Offset Correction: v_off_ini = 2777.471 uV, v_off_fin = 2782.320 uV 21:21:05 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 169.543 uVs 21:21:08 I2Y Average and RMS of Measurements VT = -159.870 +- 9.429 uVs I2Y = -14.708 +- 0.867 uTm^2 21:21:12 Wire Position: X0 = 0.0000 mm, Y0 = 0.5000 mm 21:21:16 Integrated Voltage Measurement... 21:21:25 Voltage Offset Correction: v_off_ini = 2758.811 uV, v_off_fin = 2769.760 uV 21:21:25 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -118.682 uVs 21:21:25 Integrated Voltage Measurement... 21:21:34 Voltage Offset Correction: v_off_ini = 2773.100 uV, v_off_fin = 2787.678 uV 21:21:34 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 134.324 uVs 21:21:34 Integrated Voltage Measurement... 21:21:43 Voltage Offset Correction: v_off_ini = 2811.417 uV, v_off_fin = 2780.015 uV 21:21:43 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.4999 mm dXY = -0.4999 mm VT = -80.303 uVs 21:21:43 Integrated Voltage Measurement... 21:21:52 Voltage Offset Correction: v_off_ini = 2757.831 uV, v_off_fin = 2761.453 uV 21:21:52 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = 86.356 uVs 21:21:52 Integrated Voltage Measurement... 21:22:01 Voltage Offset Correction: v_off_ini = 2786.242 uV, v_off_fin = 2795.629 uV 21:22:01 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -107.832 uVs 21:22:01 Integrated Voltage Measurement... 21:22:10 Voltage Offset Correction: v_off_ini = 2787.211 uV, v_off_fin = 2789.698 uV 21:22:10 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 121.392 uVs 21:22:14 I1X Average and RMS of Measurements VT = -108.148 +- 18.209 uVs I1X = 21.630 +- 3.641 uTm 21:22:17 Wire Position: X0 = 0.0000 mm, Y0 = 0.5000 mm 21:22:21 Integrated Voltage Measurement... 21:22:30 Voltage Offset Correction: v_off_ini = 2767.154 uV, v_off_fin = 2754.670 uV 21:22:30 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -169.158 uVs 21:22:30 Integrated Voltage Measurement... 21:22:39 Voltage Offset Correction: v_off_ini = 2776.482 uV, v_off_fin = 2783.097 uV 21:22:39 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 156.784 uVs 21:22:39 Integrated Voltage Measurement... 21:22:48 Voltage Offset Correction: v_off_ini = 2780.540 uV, v_off_fin = 2763.913 uV 21:22:48 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -137.500 uVs 21:22:48 Integrated Voltage Measurement... 21:22:57 Voltage Offset Correction: v_off_ini = 2817.176 uV, v_off_fin = 2803.106 uV 21:22:57 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 156.311 uVs 21:22:57 Integrated Voltage Measurement... 21:23:06 Voltage Offset Correction: v_off_ini = 2818.373 uV, v_off_fin = 2814.378 uV 21:23:06 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -139.361 uVs 21:23:06 Integrated Voltage Measurement... 21:23:15 Voltage Offset Correction: v_off_ini = 2803.674 uV, v_off_fin = 2818.543 uV 21:23:15 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 145.739 uVs 21:23:19 I1Y Average and RMS of Measurements VT = -150.809 +- 8.782 uVs I1Y = -30.160 +- 1.758 uTm 21:23:22 Wire Position: X0 = 0.0000 mm, Y0 = 0.5000 mm 21:23:24 Integrated Voltage Measurement... 21:23:33 Voltage Offset Correction: v_off_ini = 2767.153 uV, v_off_fin = 2802.109 uV 21:23:33 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -58.330 uVs 21:23:33 Integrated Voltage Measurement... 21:23:42 Voltage Offset Correction: v_off_ini = 2806.094 uV, v_off_fin = 2807.270 uV 21:23:42 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 79.325 uVs 21:23:42 Integrated Voltage Measurement... 21:23:51 Voltage Offset Correction: v_off_ini = 2782.915 uV, v_off_fin = 2779.094 uV 21:23:51 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -73.735 uVs 21:23:51 Integrated Voltage Measurement... 21:24:00 Voltage Offset Correction: v_off_ini = 2777.571 uV, v_off_fin = 2808.072 uV 21:24:00 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 58.569 uVs 21:24:00 Integrated Voltage Measurement... 21:24:09 Voltage Offset Correction: v_off_ini = 2800.564 uV, v_off_fin = 2777.743 uV 21:24:09 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -95.495 uVs 21:24:09 Integrated Voltage Measurement... 21:24:18 Voltage Offset Correction: v_off_ini = 2763.010 uV, v_off_fin = 2785.878 uV 21:24:18 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 92.150 uVs 21:24:21 I2X Average and RMS of Measurements VT = -76.268 +- 12.461 uVs I2X = 7.017 +- 1.146 uTm^2 21:24:23 Wire Position: X0 = 0.0000 mm, Y0 = 0.5000 mm 21:24:26 Integrated Voltage Measurement... 21:24:35 Voltage Offset Correction: v_off_ini = 2806.843 uV, v_off_fin = 2801.805 uV 21:24:35 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -140.659 uVs 21:24:35 Integrated Voltage Measurement... 21:24:44 Voltage Offset Correction: v_off_ini = 2793.303 uV, v_off_fin = 2815.086 uV 21:24:44 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 144.037 uVs 21:24:44 Integrated Voltage Measurement... 21:24:53 Voltage Offset Correction: v_off_ini = 2804.516 uV, v_off_fin = 2805.016 uV 21:24:53 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -149.645 uVs 21:24:53 Integrated Voltage Measurement... 21:25:02 Voltage Offset Correction: v_off_ini = 2823.407 uV, v_off_fin = 2812.073 uV 21:25:02 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 148.696 uVs 21:25:02 Integrated Voltage Measurement... 21:25:11 Voltage Offset Correction: v_off_ini = 2788.218 uV, v_off_fin = 2785.971 uV 21:25:11 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -172.437 uVs 21:25:11 Integrated Voltage Measurement... 21:25:20 Voltage Offset Correction: v_off_ini = 2801.208 uV, v_off_fin = 2800.875 uV 21:25:20 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 115.581 uVs 21:25:23 I2Y Average and RMS of Measurements VT = -145.176 +- 2.905 uVs I2Y = -13.356 +- 0.267 uTm^2 21:25:27 Wire Position: X0 = 0.0000 mm, Y0 = 1.0000 mm 21:25:31 Integrated Voltage Measurement... 21:25:40 Voltage Offset Correction: v_off_ini = 2843.300 uV, v_off_fin = 2816.987 uV 21:25:40 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -96.844 uVs 21:25:40 Integrated Voltage Measurement... 21:25:49 Voltage Offset Correction: v_off_ini = 2811.647 uV, v_off_fin = 2816.562 uV 21:25:49 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 97.548 uVs 21:25:49 Integrated Voltage Measurement... 21:25:58 Voltage Offset Correction: v_off_ini = 2824.775 uV, v_off_fin = 2812.498 uV 21:25:58 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -64.093 uVs 21:25:58 Integrated Voltage Measurement... 21:26:07 Voltage Offset Correction: v_off_ini = 2805.489 uV, v_off_fin = 2817.929 uV 21:26:07 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = 76.237 uVs 21:26:07 Integrated Voltage Measurement... 21:26:16 Voltage Offset Correction: v_off_ini = 2815.262 uV, v_off_fin = 2839.357 uV 21:26:16 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -62.244 uVs 21:26:16 Integrated Voltage Measurement... 21:26:25 Voltage Offset Correction: v_off_ini = 2834.280 uV, v_off_fin = 2834.196 uV 21:26:25 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 75.138 uVs 21:26:29 I1X Average and RMS of Measurements VT = -78.684 +- 13.104 uVs I1X = 15.737 +- 2.621 uTm 21:26:32 Wire Position: X0 = 0.0000 mm, Y0 = 1.0000 mm 21:26:36 Integrated Voltage Measurement... 21:26:45 Voltage Offset Correction: v_off_ini = 2817.990 uV, v_off_fin = 2829.937 uV 21:26:45 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -132.813 uVs 21:26:45 Integrated Voltage Measurement... 21:26:54 Voltage Offset Correction: v_off_ini = 2836.709 uV, v_off_fin = 2836.305 uV 21:26:54 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 145.113 uVs 21:26:54 Integrated Voltage Measurement... 21:27:03 Voltage Offset Correction: v_off_ini = 2801.812 uV, v_off_fin = 2801.735 uV 21:27:03 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -138.827 uVs 21:27:03 Integrated Voltage Measurement... 21:27:12 Voltage Offset Correction: v_off_ini = 2801.819 uV, v_off_fin = 2805.528 uV 21:27:12 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 158.738 uVs 21:27:12 Integrated Voltage Measurement... 21:27:21 Voltage Offset Correction: v_off_ini = 2820.548 uV, v_off_fin = 2818.416 uV 21:27:21 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -121.425 uVs 21:27:21 Integrated Voltage Measurement... 21:27:30 Voltage Offset Correction: v_off_ini = 2850.855 uV, v_off_fin = 2857.383 uV 21:27:30 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 122.631 uVs 21:27:34 I1Y Average and RMS of Measurements VT = -136.591 +- 11.051 uVs I1Y = -27.316 +- 2.209 uTm 21:27:36 Wire Position: X0 = 0.0000 mm, Y0 = 1.0000 mm 21:27:39 Integrated Voltage Measurement... 21:27:48 Voltage Offset Correction: v_off_ini = 2824.875 uV, v_off_fin = 2803.523 uV 21:27:48 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -51.479 uVs 21:27:48 Integrated Voltage Measurement... 21:27:57 Voltage Offset Correction: v_off_ini = 2831.342 uV, v_off_fin = 2816.538 uV 21:27:57 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 61.305 uVs 21:27:57 Integrated Voltage Measurement... 21:28:06 Voltage Offset Correction: v_off_ini = 2832.786 uV, v_off_fin = 2799.920 uV 21:28:06 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -95.646 uVs 21:28:06 Integrated Voltage Measurement... 21:28:15 Voltage Offset Correction: v_off_ini = 2830.547 uV, v_off_fin = 2819.024 uV 21:28:15 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 62.491 uVs 21:28:15 Integrated Voltage Measurement... 21:28:24 Voltage Offset Correction: v_off_ini = 2821.850 uV, v_off_fin = 2789.818 uV 21:28:24 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -115.878 uVs 21:28:24 Integrated Voltage Measurement... 21:28:33 Voltage Offset Correction: v_off_ini = 2819.271 uV, v_off_fin = 2859.947 uV 21:28:33 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 79.767 uVs 21:28:35 I2X Average and RMS of Measurements VT = -77.761 +- 16.939 uVs I2X = 7.154 +- 1.558 uTm^2 21:28:38 Wire Position: X0 = 0.0000 mm, Y0 = 1.0000 mm 21:28:41 Integrated Voltage Measurement... 21:28:50 Voltage Offset Correction: v_off_ini = 2814.779 uV, v_off_fin = 2819.977 uV 21:28:50 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -178.465 uVs 21:28:50 Integrated Voltage Measurement... 21:28:59 Voltage Offset Correction: v_off_ini = 2823.086 uV, v_off_fin = 2813.578 uV 21:28:59 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 123.154 uVs 21:28:59 Integrated Voltage Measurement... 21:29:08 Voltage Offset Correction: v_off_ini = 2821.263 uV, v_off_fin = 2818.382 uV 21:29:08 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -117.667 uVs 21:29:08 Integrated Voltage Measurement... 21:29:17 Voltage Offset Correction: v_off_ini = 2829.159 uV, v_off_fin = 2828.890 uV 21:29:17 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 123.469 uVs 21:29:17 Integrated Voltage Measurement... 21:29:26 Voltage Offset Correction: v_off_ini = 2829.331 uV, v_off_fin = 2826.283 uV 21:29:26 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -98.739 uVs 21:29:26 Integrated Voltage Measurement... 21:29:35 Voltage Offset Correction: v_off_ini = 2816.346 uV, v_off_fin = 2812.244 uV 21:29:35 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 139.756 uVs 21:29:37 I2Y Average and RMS of Measurements VT = -130.208 +- 14.577 uVs I2Y = -11.979 +- 1.341 uTm^2 21:30:09 Wire Position: X0 = -2.0000 mm, Y0 = 0.0000 mm 21:30:13 Integrated Voltage Measurement... 21:30:22 Voltage Offset Correction: v_off_ini = 2808.424 uV, v_off_fin = 2797.257 uV 21:30:22 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -86.449 uVs 21:30:22 Integrated Voltage Measurement... 21:30:31 Voltage Offset Correction: v_off_ini = 2812.977 uV, v_off_fin = 2810.514 uV 21:30:31 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 71.836 uVs 21:30:31 Integrated Voltage Measurement... 21:30:40 Voltage Offset Correction: v_off_ini = 2806.532 uV, v_off_fin = 2801.838 uV 21:30:40 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -81.387 uVs 21:30:40 Integrated Voltage Measurement... 21:30:49 Voltage Offset Correction: v_off_ini = 2801.616 uV, v_off_fin = 2791.734 uV 21:30:49 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 67.499 uVs 21:30:49 Integrated Voltage Measurement... 21:30:58 Voltage Offset Correction: v_off_ini = 2797.067 uV, v_off_fin = 2793.492 uV 21:30:58 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -93.529 uVs 21:30:58 Integrated Voltage Measurement... 21:31:07 Voltage Offset Correction: v_off_ini = 2813.476 uV, v_off_fin = 2827.109 uV 21:31:07 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = 113.839 uVs 21:31:11 I1X Average and RMS of Measurements VT = -85.757 +- 12.821 uVs I1X = 17.152 +- 2.565 uTm 21:31:13 Wire Position: X0 = -2.0000 mm, Y0 = 0.0000 mm 21:31:18 Integrated Voltage Measurement... 21:31:27 Voltage Offset Correction: v_off_ini = 2823.648 uV, v_off_fin = 2822.846 uV 21:31:27 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -164.622 uVs 21:31:27 Integrated Voltage Measurement... 21:31:36 Voltage Offset Correction: v_off_ini = 2826.023 uV, v_off_fin = 2826.033 uV 21:31:36 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 182.719 uVs 21:31:36 Integrated Voltage Measurement... 21:31:45 Voltage Offset Correction: v_off_ini = 2854.234 uV, v_off_fin = 2830.541 uV 21:31:45 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -166.735 uVs 21:31:45 Integrated Voltage Measurement... 21:31:54 Voltage Offset Correction: v_off_ini = 2842.701 uV, v_off_fin = 2868.979 uV 21:31:54 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 175.176 uVs 21:31:54 Integrated Voltage Measurement... 21:32:03 Voltage Offset Correction: v_off_ini = 2839.603 uV, v_off_fin = 2841.250 uV 21:32:03 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -161.616 uVs 21:32:03 Integrated Voltage Measurement... 21:32:12 Voltage Offset Correction: v_off_ini = 2822.390 uV, v_off_fin = 2820.036 uV 21:32:12 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 157.374 uVs 21:32:16 I1Y Average and RMS of Measurements VT = -168.040 +- 6.143 uVs I1Y = -33.608 +- 1.229 uTm 21:32:21 Wire Position: X0 = -2.0000 mm, Y0 = 0.0000 mm 21:32:23 Integrated Voltage Measurement... 21:32:32 Voltage Offset Correction: v_off_ini = 2825.305 uV, v_off_fin = 2821.583 uV 21:32:32 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -73.688 uVs 21:32:32 Integrated Voltage Measurement... 21:32:41 Voltage Offset Correction: v_off_ini = 2830.472 uV, v_off_fin = 2873.152 uV 21:32:41 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 70.332 uVs 21:32:41 Integrated Voltage Measurement... 21:32:50 Voltage Offset Correction: v_off_ini = 2848.151 uV, v_off_fin = 2819.955 uV 21:32:50 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -85.468 uVs 21:32:51 Integrated Voltage Measurement... 21:32:59 Voltage Offset Correction: v_off_ini = 2828.577 uV, v_off_fin = 2854.186 uV 21:33:00 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 93.753 uVs 21:33:00 Integrated Voltage Measurement... 21:33:09 Voltage Offset Correction: v_off_ini = 2839.467 uV, v_off_fin = 2835.745 uV 21:33:09 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -54.498 uVs 21:33:09 Integrated Voltage Measurement... 21:33:18 Voltage Offset Correction: v_off_ini = 2855.152 uV, v_off_fin = 2854.031 uV 21:33:18 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 63.663 uVs 21:33:20 I2X Average and RMS of Measurements VT = -73.567 +- 12.512 uVs I2X = 6.768 +- 1.151 uTm^2 21:33:25 Wire Position: X0 = -2.0000 mm, Y0 = 0.0000 mm 21:33:28 Integrated Voltage Measurement... 21:33:37 Voltage Offset Correction: v_off_ini = 2873.335 uV, v_off_fin = 2877.719 uV 21:33:37 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -172.998 uVs 21:33:37 Integrated Voltage Measurement... 21:33:46 Voltage Offset Correction: v_off_ini = 2872.881 uV, v_off_fin = 2871.283 uV 21:33:46 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 165.741 uVs 21:33:46 Integrated Voltage Measurement... 21:33:55 Voltage Offset Correction: v_off_ini = 2876.019 uV, v_off_fin = 2858.363 uV 21:33:55 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -145.677 uVs 21:33:55 Integrated Voltage Measurement... 21:34:04 Voltage Offset Correction: v_off_ini = 2878.383 uV, v_off_fin = 2874.712 uV 21:34:04 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 170.002 uVs 21:34:04 Integrated Voltage Measurement... 21:34:13 Voltage Offset Correction: v_off_ini = 2856.482 uV, v_off_fin = 2883.058 uV 21:34:13 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -163.340 uVs 21:34:13 Integrated Voltage Measurement... 21:34:22 Voltage Offset Correction: v_off_ini = 2908.386 uV, v_off_fin = 2887.735 uV 21:34:22 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 186.408 uVs 21:34:24 I2Y Average and RMS of Measurements VT = -167.361 +- 7.098 uVs I2Y = -15.397 +- 0.653 uTm^2 21:34:29 Wire Position: X0 = -1.0000 mm, Y0 = 0.0000 mm 21:34:33 Integrated Voltage Measurement... 21:34:42 Voltage Offset Correction: v_off_ini = 2899.113 uV, v_off_fin = 2898.633 uV 21:34:42 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -91.891 uVs 21:34:42 Integrated Voltage Measurement... 21:34:51 Voltage Offset Correction: v_off_ini = 2886.138 uV, v_off_fin = 2876.499 uV 21:34:51 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 105.116 uVs 21:34:51 Integrated Voltage Measurement... 21:35:00 Voltage Offset Correction: v_off_ini = 2878.730 uV, v_off_fin = 2880.861 uV 21:35:00 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -102.464 uVs 21:35:00 Integrated Voltage Measurement... 21:35:09 Voltage Offset Correction: v_off_ini = 2915.159 uV, v_off_fin = 2891.939 uV 21:35:09 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 55.277 uVs 21:35:09 Integrated Voltage Measurement... 21:35:18 Voltage Offset Correction: v_off_ini = 2876.058 uV, v_off_fin = 2889.169 uV 21:35:18 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -87.624 uVs 21:35:18 Integrated Voltage Measurement... 21:35:27 Voltage Offset Correction: v_off_ini = 2868.615 uV, v_off_fin = 2884.053 uV 21:35:27 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = 92.367 uVs 21:35:31 I1X Average and RMS of Measurements VT = -89.123 +- 8.039 uVs I1X = 17.826 +- 1.608 uTm 21:35:33 Wire Position: X0 = -1.0000 mm, Y0 = 0.0000 mm 21:35:38 Integrated Voltage Measurement... 21:35:47 Voltage Offset Correction: v_off_ini = 2902.119 uV, v_off_fin = 2902.612 uV 21:35:47 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -147.535 uVs 21:35:47 Integrated Voltage Measurement... 21:35:56 Voltage Offset Correction: v_off_ini = 2882.329 uV, v_off_fin = 2899.050 uV 21:35:56 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 156.973 uVs 21:35:56 Integrated Voltage Measurement... 21:36:05 Voltage Offset Correction: v_off_ini = 2896.935 uV, v_off_fin = 2869.850 uV 21:36:05 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -181.750 uVs 21:36:05 Integrated Voltage Measurement... 21:36:14 Voltage Offset Correction: v_off_ini = 2909.355 uV, v_off_fin = 2911.562 uV 21:36:14 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 165.170 uVs 21:36:14 Integrated Voltage Measurement... 21:36:23 Voltage Offset Correction: v_off_ini = 2930.379 uV, v_off_fin = 2899.922 uV 21:36:23 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -171.787 uVs 21:36:23 Integrated Voltage Measurement... 21:36:32 Voltage Offset Correction: v_off_ini = 2931.849 uV, v_off_fin = 2946.067 uV 21:36:32 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 157.099 uVs 21:36:36 I1Y Average and RMS of Measurements VT = -163.385 +- 8.690 uVs I1Y = -32.677 +- 1.738 uTm 21:36:41 Wire Position: X0 = -1.0000 mm, Y0 = 0.0000 mm 21:36:43 Integrated Voltage Measurement... 21:36:52 Voltage Offset Correction: v_off_ini = 2904.564 uV, v_off_fin = 2901.084 uV 21:36:52 Integrated Voltage Measurement dXYus = -1.0001 mm, dXYds = 0.0000 mm dXY = -1.0001 mm VT = -76.052 uVs 21:36:52 Integrated Voltage Measurement... 21:37:01 Voltage Offset Correction: v_off_ini = 2936.721 uV, v_off_fin = 2942.415 uV 21:37:01 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 47.913 uVs 21:37:01 Integrated Voltage Measurement... 21:37:10 Voltage Offset Correction: v_off_ini = 2909.157 uV, v_off_fin = 2889.940 uV 21:37:10 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -81.978 uVs 21:37:10 Integrated Voltage Measurement... 21:37:19 Voltage Offset Correction: v_off_ini = 2907.173 uV, v_off_fin = 2920.005 uV 21:37:19 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 71.119 uVs 21:37:20 Integrated Voltage Measurement... 21:37:28 Voltage Offset Correction: v_off_ini = 2925.813 uV, v_off_fin = 2922.868 uV 21:37:28 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -70.621 uVs 21:37:29 Integrated Voltage Measurement... 21:37:37 Voltage Offset Correction: v_off_ini = 2917.848 uV, v_off_fin = 2929.991 uV 21:37:37 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 80.044 uVs 21:37:40 I2X Average and RMS of Measurements VT = -71.288 +- 6.599 uVs I2X = 6.558 +- 0.607 uTm^2 21:37:45 Wire Position: X0 = -1.0000 mm, Y0 = 0.0000 mm 21:37:48 Integrated Voltage Measurement... 21:37:56 Voltage Offset Correction: v_off_ini = 2921.195 uV, v_off_fin = 2911.273 uV 21:37:57 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -189.197 uVs 21:37:57 Integrated Voltage Measurement... 21:38:06 Voltage Offset Correction: v_off_ini = 2923.158 uV, v_off_fin = 2939.587 uV 21:38:06 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 180.172 uVs 21:38:06 Integrated Voltage Measurement... 21:38:15 Voltage Offset Correction: v_off_ini = 2943.459 uV, v_off_fin = 2932.898 uV 21:38:15 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -163.914 uVs 21:38:15 Integrated Voltage Measurement... 21:38:24 Voltage Offset Correction: v_off_ini = 2929.645 uV, v_off_fin = 2944.115 uV 21:38:24 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 183.996 uVs 21:38:24 Integrated Voltage Measurement... 21:38:33 Voltage Offset Correction: v_off_ini = 2933.027 uV, v_off_fin = 2905.821 uV 21:38:33 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -156.056 uVs 21:38:33 Integrated Voltage Measurement... 21:38:42 Voltage Offset Correction: v_off_ini = 2916.285 uV, v_off_fin = 2923.708 uV 21:38:42 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 161.209 uVs 21:38:44 I2Y Average and RMS of Measurements VT = -172.424 +- 10.690 uVs I2Y = -15.863 +- 0.984 uTm^2 21:38:49 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 21:38:53 Integrated Voltage Measurement... 21:39:02 Voltage Offset Correction: v_off_ini = 2904.059 uV, v_off_fin = 2903.869 uV 21:39:02 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -87.892 uVs 21:39:02 Integrated Voltage Measurement... 21:39:11 Voltage Offset Correction: v_off_ini = 2919.993 uV, v_off_fin = 2912.200 uV 21:39:11 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 59.362 uVs 21:39:11 Integrated Voltage Measurement... 21:39:20 Voltage Offset Correction: v_off_ini = 2916.239 uV, v_off_fin = 2905.667 uV 21:39:20 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -67.959 uVs 21:39:20 Integrated Voltage Measurement... 21:39:29 Voltage Offset Correction: v_off_ini = 2904.408 uV, v_off_fin = 2912.648 uV 21:39:29 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 85.513 uVs 21:39:29 Integrated Voltage Measurement... 21:39:38 Voltage Offset Correction: v_off_ini = 2942.761 uV, v_off_fin = 2911.064 uV 21:39:38 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -79.541 uVs 21:39:38 Integrated Voltage Measurement... 21:39:47 Voltage Offset Correction: v_off_ini = 2922.833 uV, v_off_fin = 2932.369 uV 21:39:47 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 73.708 uVs 21:39:51 I1X Average and RMS of Measurements VT = -75.663 +- 1.440 uVs I1X = 15.133 +- 0.288 uTm 21:39:54 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 21:39:58 Integrated Voltage Measurement... 21:40:07 Voltage Offset Correction: v_off_ini = 2902.922 uV, v_off_fin = 2896.134 uV 21:40:07 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -171.047 uVs 21:40:07 Integrated Voltage Measurement... 21:40:16 Voltage Offset Correction: v_off_ini = 2893.246 uV, v_off_fin = 2918.292 uV 21:40:16 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 160.336 uVs 21:40:16 Integrated Voltage Measurement... 21:40:25 Voltage Offset Correction: v_off_ini = 2914.639 uV, v_off_fin = 2903.656 uV 21:40:25 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -168.542 uVs 21:40:25 Integrated Voltage Measurement... 21:40:34 Voltage Offset Correction: v_off_ini = 2906.811 uV, v_off_fin = 2899.355 uV 21:40:34 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 168.088 uVs 21:40:34 Integrated Voltage Measurement... 21:40:43 Voltage Offset Correction: v_off_ini = 2904.660 uV, v_off_fin = 2891.085 uV 21:40:43 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -170.226 uVs 21:40:43 Integrated Voltage Measurement... 21:40:52 Voltage Offset Correction: v_off_ini = 2911.755 uV, v_off_fin = 2900.707 uV 21:40:52 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 168.047 uVs 21:40:56 I1Y Average and RMS of Measurements VT = -167.714 +- 1.469 uVs I1Y = -33.543 +- 0.294 uTm 21:41:01 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 21:41:04 Integrated Voltage Measurement... 21:41:13 Voltage Offset Correction: v_off_ini = 2913.780 uV, v_off_fin = 2895.704 uV 21:41:13 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -36.706 uVs 21:41:13 Integrated Voltage Measurement... 21:41:22 Voltage Offset Correction: v_off_ini = 2898.203 uV, v_off_fin = 2912.847 uV 21:41:22 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 38.932 uVs 21:41:22 Integrated Voltage Measurement... 21:41:31 Voltage Offset Correction: v_off_ini = 2924.020 uV, v_off_fin = 2894.261 uV 21:41:31 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -97.805 uVs 21:41:31 Integrated Voltage Measurement... 21:41:40 Voltage Offset Correction: v_off_ini = 2920.935 uV, v_off_fin = 2924.040 uV 21:41:40 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 79.626 uVs 21:41:40 Integrated Voltage Measurement... 21:41:49 Voltage Offset Correction: v_off_ini = 2913.089 uV, v_off_fin = 2929.795 uV 21:41:49 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -76.380 uVs 21:41:49 Integrated Voltage Measurement... 21:41:58 Voltage Offset Correction: v_off_ini = 2926.458 uV, v_off_fin = 2943.349 uV 21:41:58 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 77.619 uVs 21:42:00 I2X Average and RMS of Measurements VT = -67.845 +- 21.763 uVs I2X = 6.242 +- 2.002 uTm^2 21:42:05 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 21:42:08 Integrated Voltage Measurement... 21:42:17 Voltage Offset Correction: v_off_ini = 2920.888 uV, v_off_fin = 2921.075 uV 21:42:17 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -159.675 uVs 21:42:17 Integrated Voltage Measurement... 21:42:26 Voltage Offset Correction: v_off_ini = 2911.375 uV, v_off_fin = 2915.383 uV 21:42:26 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 167.656 uVs 21:42:26 Integrated Voltage Measurement... 21:42:35 Voltage Offset Correction: v_off_ini = 2915.737 uV, v_off_fin = 2904.314 uV 21:42:35 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -172.064 uVs 21:42:35 Integrated Voltage Measurement... 21:42:44 Voltage Offset Correction: v_off_ini = 2901.159 uV, v_off_fin = 2899.812 uV 21:42:44 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 149.789 uVs 21:42:44 Integrated Voltage Measurement... 21:42:53 Voltage Offset Correction: v_off_ini = 2910.673 uV, v_off_fin = 2913.161 uV 21:42:53 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -155.332 uVs 21:42:53 Integrated Voltage Measurement... 21:43:02 Voltage Offset Correction: v_off_ini = 2910.042 uV, v_off_fin = 2917.561 uV 21:43:02 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 147.426 uVs 21:43:04 I2Y Average and RMS of Measurements VT = -158.657 +- 5.267 uVs I2Y = -14.596 +- 0.485 uTm^2 21:43:09 Wire Position: X0 = 1.0000 mm, Y0 = 0.0000 mm 21:43:13 Integrated Voltage Measurement... 21:43:22 Voltage Offset Correction: v_off_ini = 2904.490 uV, v_off_fin = 2891.198 uV 21:43:22 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.4998 mm dXY = -0.4999 mm VT = -70.681 uVs 21:43:22 Integrated Voltage Measurement... 21:43:31 Voltage Offset Correction: v_off_ini = 2927.198 uV, v_off_fin = 2926.444 uV 21:43:31 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 73.811 uVs 21:43:31 Integrated Voltage Measurement... 21:43:40 Voltage Offset Correction: v_off_ini = 2918.558 uV, v_off_fin = 2915.748 uV 21:43:40 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -87.342 uVs 21:43:40 Integrated Voltage Measurement... 21:43:49 Voltage Offset Correction: v_off_ini = 2917.387 uV, v_off_fin = 2911.127 uV 21:43:49 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 80.564 uVs 21:43:49 Integrated Voltage Measurement... 21:43:58 Voltage Offset Correction: v_off_ini = 2897.138 uV, v_off_fin = 2907.986 uV 21:43:58 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -98.017 uVs 21:43:58 Integrated Voltage Measurement... 21:44:07 Voltage Offset Correction: v_off_ini = 2921.382 uV, v_off_fin = 2918.963 uV 21:44:07 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 56.040 uVs 21:44:11 I1X Average and RMS of Measurements VT = -77.743 +- 4.806 uVs I1X = 15.549 +- 0.961 uTm 21:44:14 Wire Position: X0 = 1.0000 mm, Y0 = 0.0000 mm 21:44:18 Integrated Voltage Measurement... 21:44:27 Voltage Offset Correction: v_off_ini = 2917.134 uV, v_off_fin = 2898.907 uV 21:44:27 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -158.066 uVs 21:44:27 Integrated Voltage Measurement... 21:44:36 Voltage Offset Correction: v_off_ini = 2912.652 uV, v_off_fin = 2927.922 uV 21:44:36 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 156.578 uVs 21:44:36 Integrated Voltage Measurement... 21:44:45 Voltage Offset Correction: v_off_ini = 2929.155 uV, v_off_fin = 2930.072 uV 21:44:45 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -161.017 uVs 21:44:45 Integrated Voltage Measurement... 21:44:54 Voltage Offset Correction: v_off_ini = 2938.251 uV, v_off_fin = 2933.307 uV 21:44:54 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 160.400 uVs 21:44:54 Integrated Voltage Measurement... 21:45:03 Voltage Offset Correction: v_off_ini = 2931.103 uV, v_off_fin = 2926.530 uV 21:45:03 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -175.426 uVs 21:45:03 Integrated Voltage Measurement... 21:45:12 Voltage Offset Correction: v_off_ini = 2932.998 uV, v_off_fin = 2949.846 uV 21:45:12 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 162.469 uVs 21:45:16 I1Y Average and RMS of Measurements VT = -162.326 +- 4.882 uVs I1Y = -32.466 +- 0.976 uTm 21:45:21 Wire Position: X0 = 1.0000 mm, Y0 = 0.0000 mm 21:45:24 Integrated Voltage Measurement... 21:45:33 Voltage Offset Correction: v_off_ini = 2956.893 uV, v_off_fin = 2932.698 uV 21:45:33 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -86.273 uVs 21:45:33 Integrated Voltage Measurement... 21:45:42 Voltage Offset Correction: v_off_ini = 2947.734 uV, v_off_fin = 2959.657 uV 21:45:42 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 67.396 uVs 21:45:42 Integrated Voltage Measurement... 21:45:51 Voltage Offset Correction: v_off_ini = 2936.199 uV, v_off_fin = 2930.183 uV 21:45:51 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -56.162 uVs 21:45:51 Integrated Voltage Measurement... 21:46:00 Voltage Offset Correction: v_off_ini = 2963.810 uV, v_off_fin = 2978.110 uV 21:46:00 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 60.292 uVs 21:46:00 Integrated Voltage Measurement... 21:46:09 Voltage Offset Correction: v_off_ini = 2946.616 uV, v_off_fin = 2935.482 uV 21:46:09 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -84.634 uVs 21:46:09 Integrated Voltage Measurement... 21:46:18 Voltage Offset Correction: v_off_ini = 2946.661 uV, v_off_fin = 2962.849 uV 21:46:18 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 91.644 uVs 21:46:20 I2X Average and RMS of Measurements VT = -74.400 +- 12.332 uVs I2X = 6.845 +- 1.135 uTm^2 21:46:25 Wire Position: X0 = 1.0000 mm, Y0 = 0.0000 mm 21:46:28 Integrated Voltage Measurement... 21:46:37 Voltage Offset Correction: v_off_ini = 2942.428 uV, v_off_fin = 2950.132 uV 21:46:37 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -168.068 uVs 21:46:37 Integrated Voltage Measurement... 21:46:46 Voltage Offset Correction: v_off_ini = 2948.261 uV, v_off_fin = 2961.779 uV 21:46:46 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 156.760 uVs 21:46:46 Integrated Voltage Measurement... 21:46:55 Voltage Offset Correction: v_off_ini = 2939.849 uV, v_off_fin = 2946.520 uV 21:46:55 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -178.333 uVs 21:46:55 Integrated Voltage Measurement... 21:47:04 Voltage Offset Correction: v_off_ini = 2928.330 uV, v_off_fin = 2930.391 uV 21:47:04 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 172.070 uVs 21:47:04 Integrated Voltage Measurement... 21:47:13 Voltage Offset Correction: v_off_ini = 2949.016 uV, v_off_fin = 2940.408 uV 21:47:13 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -153.243 uVs 21:47:13 Integrated Voltage Measurement... 21:47:22 Voltage Offset Correction: v_off_ini = 2949.745 uV, v_off_fin = 2941.237 uV 21:47:22 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 143.575 uVs 21:47:25 I2Y Average and RMS of Measurements VT = -162.008 +- 10.942 uVs I2Y = -14.905 +- 1.007 uTm^2 21:47:29 Wire Position: X0 = 2.0000 mm, Y0 = 0.0000 mm 21:47:34 Integrated Voltage Measurement... 21:47:42 Voltage Offset Correction: v_off_ini = 2950.851 uV, v_off_fin = 2941.392 uV 21:47:43 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.4998 mm dXY = -0.4999 mm VT = -71.203 uVs 21:47:43 Integrated Voltage Measurement... 21:47:51 Voltage Offset Correction: v_off_ini = 2961.596 uV, v_off_fin = 2982.511 uV 21:47:52 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 81.142 uVs 21:47:52 Integrated Voltage Measurement... 21:48:00 Voltage Offset Correction: v_off_ini = 2963.385 uV, v_off_fin = 2953.035 uV 21:48:01 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -79.584 uVs 21:48:01 Integrated Voltage Measurement... 21:48:09 Voltage Offset Correction: v_off_ini = 2949.574 uV, v_off_fin = 2973.588 uV 21:48:10 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 83.653 uVs 21:48:10 Integrated Voltage Measurement... 21:48:18 Voltage Offset Correction: v_off_ini = 2960.891 uV, v_off_fin = 2981.491 uV 21:48:19 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -104.283 uVs 21:48:19 Integrated Voltage Measurement... 21:48:28 Voltage Offset Correction: v_off_ini = 2978.557 uV, v_off_fin = 2985.223 uV 21:48:28 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 93.937 uVs 21:48:32 I1X Average and RMS of Measurements VT = -85.634 +- 9.785 uVs I1X = 17.127 +- 1.956 uTm 21:48:34 Wire Position: X0 = 2.0000 mm, Y0 = 0.0000 mm 21:48:38 Integrated Voltage Measurement... 21:48:47 Voltage Offset Correction: v_off_ini = 2982.196 uV, v_off_fin = 2971.942 uV 21:48:47 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -189.206 uVs 21:48:47 Integrated Voltage Measurement... 21:48:56 Voltage Offset Correction: v_off_ini = 2988.059 uV, v_off_fin = 2990.448 uV 21:48:56 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 171.075 uVs 21:48:56 Integrated Voltage Measurement... 21:49:05 Voltage Offset Correction: v_off_ini = 2970.341 uV, v_off_fin = 2962.875 uV 21:49:05 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -168.452 uVs 21:49:05 Integrated Voltage Measurement... 21:49:14 Voltage Offset Correction: v_off_ini = 2968.462 uV, v_off_fin = 2975.372 uV 21:49:14 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 168.029 uVs 21:49:14 Integrated Voltage Measurement... 21:49:23 Voltage Offset Correction: v_off_ini = 2958.750 uV, v_off_fin = 2953.613 uV 21:49:23 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -172.683 uVs 21:49:23 Integrated Voltage Measurement... 21:49:32 Voltage Offset Correction: v_off_ini = 2978.891 uV, v_off_fin = 2982.242 uV 21:49:32 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 179.432 uVs 21:49:37 I1Y Average and RMS of Measurements VT = -174.813 +- 4.937 uVs I1Y = -34.963 +- 0.987 uTm 21:49:42 Wire Position: X0 = 2.0000 mm, Y0 = 0.0000 mm 21:49:44 Integrated Voltage Measurement... 21:49:53 Voltage Offset Correction: v_off_ini = 2971.762 uV, v_off_fin = 2958.206 uV 21:49:53 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -71.065 uVs 21:49:53 Integrated Voltage Measurement... 21:50:02 Voltage Offset Correction: v_off_ini = 2958.964 uV, v_off_fin = 2972.571 uV 21:50:02 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 51.750 uVs 21:50:02 Integrated Voltage Measurement... 21:50:11 Voltage Offset Correction: v_off_ini = 2957.899 uV, v_off_fin = 2951.783 uV 21:50:11 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -63.069 uVs 21:50:11 Integrated Voltage Measurement... 21:50:20 Voltage Offset Correction: v_off_ini = 2951.546 uV, v_off_fin = 2955.920 uV 21:50:20 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 68.206 uVs 21:50:20 Integrated Voltage Measurement... 21:50:29 Voltage Offset Correction: v_off_ini = 2953.020 uV, v_off_fin = 2936.559 uV 21:50:29 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -46.894 uVs 21:50:29 Integrated Voltage Measurement... 21:50:38 Voltage Offset Correction: v_off_ini = 2960.990 uV, v_off_fin = 2960.863 uV 21:50:38 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 51.118 uVs 21:50:41 I2X Average and RMS of Measurements VT = -58.684 +- 7.058 uVs I2X = 5.399 +- 0.649 uTm^2 21:50:46 Wire Position: X0 = 2.0000 mm, Y0 = 0.0000 mm 21:50:48 Integrated Voltage Measurement... 21:50:57 Voltage Offset Correction: v_off_ini = 2932.199 uV, v_off_fin = 2941.916 uV 21:50:57 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -143.388 uVs 21:50:57 Integrated Voltage Measurement... 21:51:06 Voltage Offset Correction: v_off_ini = 2952.477 uV, v_off_fin = 2993.119 uV 21:51:06 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 129.383 uVs 21:51:06 Integrated Voltage Measurement... 21:51:15 Voltage Offset Correction: v_off_ini = 2980.851 uV, v_off_fin = 2959.497 uV 21:51:16 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -176.760 uVs 21:51:16 Integrated Voltage Measurement... 21:51:24 Voltage Offset Correction: v_off_ini = 2967.596 uV, v_off_fin = 2957.842 uV 21:51:25 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 173.623 uVs 21:51:25 Integrated Voltage Measurement... 21:51:34 Voltage Offset Correction: v_off_ini = 2957.546 uV, v_off_fin = 2974.605 uV 21:51:34 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -163.126 uVs 21:51:34 Integrated Voltage Measurement... 21:51:43 Voltage Offset Correction: v_off_ini = 2977.813 uV, v_off_fin = 2979.344 uV 21:51:43 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 188.921 uVs 21:51:45 I2Y Average and RMS of Measurements VT = -162.533 +- 18.492 uVs I2Y = -14.953 +- 1.701 uTm^2 21:51:53 Wire Position: X0 = 0.0000 mm, Y0 = -1.0000 mm 21:51:57 Integrated Voltage Measurement... 21:52:06 Voltage Offset Correction: v_off_ini = 2948.546 uV, v_off_fin = 2967.154 uV 21:52:06 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -61.455 uVs 21:52:06 Integrated Voltage Measurement... 21:52:15 Voltage Offset Correction: v_off_ini = 2966.332 uV, v_off_fin = 2946.568 uV 21:52:15 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 90.886 uVs 21:52:15 Integrated Voltage Measurement... 21:52:24 Voltage Offset Correction: v_off_ini = 2970.356 uV, v_off_fin = 2966.716 uV 21:52:24 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -91.251 uVs 21:52:24 Integrated Voltage Measurement... 21:52:33 Voltage Offset Correction: v_off_ini = 2978.559 uV, v_off_fin = 2961.066 uV 21:52:33 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 83.334 uVs 21:52:33 Integrated Voltage Measurement... 21:52:42 Voltage Offset Correction: v_off_ini = 2954.224 uV, v_off_fin = 2980.265 uV 21:52:42 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -60.307 uVs 21:52:42 Integrated Voltage Measurement... 21:52:51 Voltage Offset Correction: v_off_ini = 2953.719 uV, v_off_fin = 2974.827 uV 21:52:51 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 92.287 uVs 21:52:55 I1X Average and RMS of Measurements VT = -79.920 +- 5.213 uVs I1X = 15.984 +- 1.043 uTm 21:52:58 Wire Position: X0 = 0.0000 mm, Y0 = -1.0000 mm 21:53:02 Integrated Voltage Measurement... 21:53:11 Voltage Offset Correction: v_off_ini = 2966.102 uV, v_off_fin = 2961.297 uV 21:53:11 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -185.548 uVs 21:53:11 Integrated Voltage Measurement... 21:53:20 Voltage Offset Correction: v_off_ini = 2966.936 uV, v_off_fin = 2975.106 uV 21:53:20 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 159.715 uVs 21:53:20 Integrated Voltage Measurement... 21:53:29 Voltage Offset Correction: v_off_ini = 2950.368 uV, v_off_fin = 2956.811 uV 21:53:29 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -171.834 uVs 21:53:29 Integrated Voltage Measurement... 21:53:38 Voltage Offset Correction: v_off_ini = 2956.325 uV, v_off_fin = 2972.662 uV 21:53:38 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 163.922 uVs 21:53:38 Integrated Voltage Measurement... 21:53:47 Voltage Offset Correction: v_off_ini = 2965.229 uV, v_off_fin = 2949.701 uV 21:53:47 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -155.169 uVs 21:53:47 Integrated Voltage Measurement... 21:53:56 Voltage Offset Correction: v_off_ini = 2965.020 uV, v_off_fin = 2943.493 uV 21:53:56 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 164.330 uVs 21:54:00 I1Y Average and RMS of Measurements VT = -166.753 +- 5.319 uVs I1Y = -33.351 +- 1.064 uTm 21:54:03 Wire Position: X0 = 0.0000 mm, Y0 = -1.0000 mm 21:54:05 Integrated Voltage Measurement... 21:54:14 Voltage Offset Correction: v_off_ini = 2935.908 uV, v_off_fin = 2937.251 uV 21:54:14 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -84.165 uVs 21:54:14 Integrated Voltage Measurement... 21:54:23 Voltage Offset Correction: v_off_ini = 2942.266 uV, v_off_fin = 2933.492 uV 21:54:23 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 105.939 uVs 21:54:23 Integrated Voltage Measurement... 21:54:32 Voltage Offset Correction: v_off_ini = 2933.674 uV, v_off_fin = 2914.170 uV 21:54:32 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -61.038 uVs 21:54:32 Integrated Voltage Measurement... 21:54:41 Voltage Offset Correction: v_off_ini = 2921.755 uV, v_off_fin = 2932.637 uV 21:54:41 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 97.667 uVs 21:54:41 Integrated Voltage Measurement... 21:54:50 Voltage Offset Correction: v_off_ini = 2944.148 uV, v_off_fin = 2913.237 uV 21:54:50 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -63.508 uVs 21:54:50 Integrated Voltage Measurement... 21:54:59 Voltage Offset Correction: v_off_ini = 2934.506 uV, v_off_fin = 2967.202 uV 21:54:59 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 84.581 uVs 21:55:02 I2X Average and RMS of Measurements VT = -82.816 +- 8.919 uVs I2X = 7.619 +- 0.821 uTm^2 21:55:04 Wire Position: X0 = 0.0000 mm, Y0 = -1.0000 mm 21:55:07 Integrated Voltage Measurement... 21:55:16 Voltage Offset Correction: v_off_ini = 2929.831 uV, v_off_fin = 2922.285 uV 21:55:16 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -168.932 uVs 21:55:16 Integrated Voltage Measurement... 21:55:25 Voltage Offset Correction: v_off_ini = 2938.093 uV, v_off_fin = 2949.938 uV 21:55:25 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 168.688 uVs 21:55:25 Integrated Voltage Measurement... 21:55:34 Voltage Offset Correction: v_off_ini = 2937.952 uV, v_off_fin = 2932.523 uV 21:55:34 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -178.662 uVs 21:55:34 Integrated Voltage Measurement... 21:55:43 Voltage Offset Correction: v_off_ini = 2928.082 uV, v_off_fin = 2930.311 uV 21:55:43 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 159.621 uVs 21:55:43 Integrated Voltage Measurement... 21:55:52 Voltage Offset Correction: v_off_ini = 2947.712 uV, v_off_fin = 2925.895 uV 21:55:52 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -161.967 uVs 21:55:52 Integrated Voltage Measurement... 21:56:01 Voltage Offset Correction: v_off_ini = 2934.142 uV, v_off_fin = 2919.912 uV 21:56:01 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 146.114 uVs 21:56:03 I2Y Average and RMS of Measurements VT = -163.997 +- 7.042 uVs I2Y = -15.088 +- 0.648 uTm^2 21:56:07 Wire Position: X0 = 0.0000 mm, Y0 = -0.5000 mm 21:56:12 Integrated Voltage Measurement... 21:56:21 Voltage Offset Correction: v_off_ini = 2921.412 uV, v_off_fin = 2924.900 uV 21:56:21 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -93.043 uVs 21:56:21 Integrated Voltage Measurement... 21:56:30 Voltage Offset Correction: v_off_ini = 2918.431 uV, v_off_fin = 2933.937 uV 21:56:30 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 101.254 uVs 21:56:30 Integrated Voltage Measurement... 21:56:39 Voltage Offset Correction: v_off_ini = 2928.081 uV, v_off_fin = 2933.252 uV 21:56:39 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -56.599 uVs 21:56:39 Integrated Voltage Measurement... 21:56:48 Voltage Offset Correction: v_off_ini = 2918.744 uV, v_off_fin = 2937.227 uV 21:56:48 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 68.928 uVs 21:56:48 Integrated Voltage Measurement... 21:56:57 Voltage Offset Correction: v_off_ini = 2950.054 uV, v_off_fin = 2933.225 uV 21:56:57 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -99.028 uVs 21:56:57 Integrated Voltage Measurement... 21:57:06 Voltage Offset Correction: v_off_ini = 2919.357 uV, v_off_fin = 2925.862 uV 21:57:06 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 76.367 uVs 21:57:10 I1X Average and RMS of Measurements VT = -82.537 +- 14.504 uVs I1X = 16.508 +- 2.901 uTm 21:57:12 Wire Position: X0 = 0.0000 mm, Y0 = -0.5000 mm 21:57:17 Integrated Voltage Measurement... 21:57:26 Voltage Offset Correction: v_off_ini = 2901.037 uV, v_off_fin = 2891.409 uV 21:57:26 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -167.899 uVs 21:57:26 Integrated Voltage Measurement... 21:57:35 Voltage Offset Correction: v_off_ini = 2904.039 uV, v_off_fin = 2940.286 uV 21:57:35 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 164.064 uVs 21:57:35 Integrated Voltage Measurement... 21:57:44 Voltage Offset Correction: v_off_ini = 2955.703 uV, v_off_fin = 2923.302 uV 21:57:44 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -151.929 uVs 21:57:44 Integrated Voltage Measurement... 21:57:53 Voltage Offset Correction: v_off_ini = 2908.239 uV, v_off_fin = 2913.679 uV 21:57:53 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 170.783 uVs 21:57:53 Integrated Voltage Measurement... 21:58:02 Voltage Offset Correction: v_off_ini = 2920.969 uV, v_off_fin = 2945.384 uV 21:58:02 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -174.657 uVs 21:58:02 Integrated Voltage Measurement... 21:58:11 Voltage Offset Correction: v_off_ini = 2923.959 uV, v_off_fin = 2942.800 uV 21:58:11 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 179.819 uVs 21:58:15 I1Y Average and RMS of Measurements VT = -168.192 +- 6.670 uVs I1Y = -33.638 +- 1.334 uTm 21:58:17 Wire Position: X0 = 0.0000 mm, Y0 = -0.5000 mm 21:58:20 Integrated Voltage Measurement... 21:58:29 Voltage Offset Correction: v_off_ini = 2920.363 uV, v_off_fin = 2915.040 uV 21:58:29 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -79.899 uVs 21:58:29 Integrated Voltage Measurement... 21:58:38 Voltage Offset Correction: v_off_ini = 2901.547 uV, v_off_fin = 2925.096 uV 21:58:38 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 62.722 uVs 21:58:38 Integrated Voltage Measurement... 21:58:47 Voltage Offset Correction: v_off_ini = 2921.370 uV, v_off_fin = 2908.675 uV 21:58:47 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -77.257 uVs 21:58:47 Integrated Voltage Measurement... 21:58:56 Voltage Offset Correction: v_off_ini = 2922.035 uV, v_off_fin = 2944.133 uV 21:58:56 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 62.657 uVs 21:58:56 Integrated Voltage Measurement... 21:59:05 Voltage Offset Correction: v_off_ini = 2947.430 uV, v_off_fin = 2925.179 uV 21:59:05 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -81.892 uVs 21:59:05 Integrated Voltage Measurement... 21:59:14 Voltage Offset Correction: v_off_ini = 2933.299 uV, v_off_fin = 2941.613 uV 21:59:14 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 53.682 uVs 21:59:16 I2X Average and RMS of Measurements VT = -69.685 +- 1.451 uVs I2X = 6.411 +- 0.134 uTm^2 21:59:19 Wire Position: X0 = 0.0000 mm, Y0 = -0.5000 mm 21:59:21 Integrated Voltage Measurement... 21:59:30 Voltage Offset Correction: v_off_ini = 2942.630 uV, v_off_fin = 2950.434 uV 21:59:30 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -155.380 uVs 21:59:30 Integrated Voltage Measurement... 21:59:39 Voltage Offset Correction: v_off_ini = 2949.553 uV, v_off_fin = 2952.818 uV 21:59:39 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 150.384 uVs 21:59:40 Integrated Voltage Measurement... 21:59:49 Voltage Offset Correction: v_off_ini = 2954.943 uV, v_off_fin = 2934.178 uV 21:59:49 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -183.664 uVs 21:59:49 Integrated Voltage Measurement... 21:59:58 Voltage Offset Correction: v_off_ini = 2930.346 uV, v_off_fin = 2944.914 uV 21:59:58 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 175.782 uVs 21:59:58 Integrated Voltage Measurement... 22:00:07 Voltage Offset Correction: v_off_ini = 2950.462 uV, v_off_fin = 2934.809 uV 22:00:07 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -162.967 uVs 22:00:07 Integrated Voltage Measurement... 22:00:16 Voltage Offset Correction: v_off_ini = 2953.791 uV, v_off_fin = 2944.694 uV 22:00:16 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 183.776 uVs 22:00:18 I2Y Average and RMS of Measurements VT = -168.659 +- 11.453 uVs I2Y = -15.517 +- 1.054 uTm^2 22:00:22 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 22:00:26 Integrated Voltage Measurement... 22:00:35 Voltage Offset Correction: v_off_ini = 2927.446 uV, v_off_fin = 2923.261 uV 22:00:35 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -64.874 uVs 22:00:35 Integrated Voltage Measurement... 22:00:44 Voltage Offset Correction: v_off_ini = 2919.556 uV, v_off_fin = 2924.148 uV 22:00:44 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 95.623 uVs 22:00:44 Integrated Voltage Measurement... 22:00:53 Voltage Offset Correction: v_off_ini = 2950.197 uV, v_off_fin = 2942.134 uV 22:00:53 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -76.389 uVs 22:00:53 Integrated Voltage Measurement... 22:01:02 Voltage Offset Correction: v_off_ini = 2954.350 uV, v_off_fin = 2925.304 uV 22:01:02 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 50.587 uVs 22:01:02 Integrated Voltage Measurement... 22:01:11 Voltage Offset Correction: v_off_ini = 2928.631 uV, v_off_fin = 2950.028 uV 22:01:11 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -96.221 uVs 22:01:11 Integrated Voltage Measurement... 22:01:20 Voltage Offset Correction: v_off_ini = 2950.708 uV, v_off_fin = 2967.178 uV 22:01:20 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 77.773 uVs 22:01:25 I1X Average and RMS of Measurements VT = -76.911 +- 9.883 uVs I1X = 15.382 +- 1.977 uTm 22:01:27 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 22:01:31 Integrated Voltage Measurement... 22:01:40 Voltage Offset Correction: v_off_ini = 2935.501 uV, v_off_fin = 2928.534 uV 22:01:40 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -152.705 uVs 22:01:40 Integrated Voltage Measurement... 22:01:49 Voltage Offset Correction: v_off_ini = 2933.468 uV, v_off_fin = 2948.951 uV 22:01:49 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 172.680 uVs 22:01:49 Integrated Voltage Measurement... 22:01:58 Voltage Offset Correction: v_off_ini = 2944.047 uV, v_off_fin = 2935.689 uV 22:01:58 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -157.549 uVs 22:01:58 Integrated Voltage Measurement... 22:02:07 Voltage Offset Correction: v_off_ini = 2965.911 uV, v_off_fin = 2959.223 uV 22:02:07 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 166.110 uVs 22:02:07 Integrated Voltage Measurement... 22:02:16 Voltage Offset Correction: v_off_ini = 2942.993 uV, v_off_fin = 2954.809 uV 22:02:16 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -155.817 uVs 22:02:16 Integrated Voltage Measurement... 22:02:25 Voltage Offset Correction: v_off_ini = 2932.281 uV, v_off_fin = 2948.106 uV 22:02:25 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 163.988 uVs 22:02:29 I1Y Average and RMS of Measurements VT = -161.475 +- 1.166 uVs I1Y = -32.295 +- 0.233 uTm 22:02:32 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 22:02:35 Integrated Voltage Measurement... 22:02:44 Voltage Offset Correction: v_off_ini = 2973.873 uV, v_off_fin = 2954.097 uV 22:02:44 Integrated Voltage Measurement dXYus = -1.0001 mm, dXYds = 0.0000 mm dXY = -1.0001 mm VT = -126.692 uVs 22:02:44 Integrated Voltage Measurement... 22:02:53 Voltage Offset Correction: v_off_ini = 2954.817 uV, v_off_fin = 2974.222 uV 22:02:53 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 76.729 uVs 22:02:53 Integrated Voltage Measurement... 22:03:02 Voltage Offset Correction: v_off_ini = 2958.490 uV, v_off_fin = 2944.069 uV 22:03:02 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -90.475 uVs 22:03:02 Integrated Voltage Measurement... 22:03:11 Voltage Offset Correction: v_off_ini = 2955.864 uV, v_off_fin = 2959.881 uV 22:03:11 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 83.571 uVs 22:03:11 Integrated Voltage Measurement... 22:03:20 Voltage Offset Correction: v_off_ini = 2951.121 uV, v_off_fin = 2932.136 uV 22:03:20 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -64.368 uVs 22:03:20 Integrated Voltage Measurement... 22:03:29 Voltage Offset Correction: v_off_ini = 2956.386 uV, v_off_fin = 2985.855 uV 22:03:29 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 69.076 uVs 22:03:31 I2X Average and RMS of Measurements VT = -85.152 +- 14.345 uVs I2X = 7.834 +- 1.320 uTm^2 22:03:34 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 22:03:36 Integrated Voltage Measurement... 22:03:45 Voltage Offset Correction: v_off_ini = 2996.547 uV, v_off_fin = 2959.366 uV 22:03:45 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -185.017 uVs 22:03:45 Integrated Voltage Measurement... 22:03:54 Voltage Offset Correction: v_off_ini = 2982.988 uV, v_off_fin = 2980.039 uV 22:03:54 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 154.939 uVs 22:03:54 Integrated Voltage Measurement... 22:04:03 Voltage Offset Correction: v_off_ini = 2963.918 uV, v_off_fin = 2966.815 uV 22:04:03 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -164.326 uVs 22:04:03 Integrated Voltage Measurement... 22:04:12 Voltage Offset Correction: v_off_ini = 2961.386 uV, v_off_fin = 2975.482 uV 22:04:12 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 169.665 uVs 22:04:12 Integrated Voltage Measurement... 22:04:21 Voltage Offset Correction: v_off_ini = 2998.143 uV, v_off_fin = 2990.142 uV 22:04:21 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -147.393 uVs 22:04:21 Integrated Voltage Measurement... 22:04:30 Voltage Offset Correction: v_off_ini = 2954.457 uV, v_off_fin = 2974.216 uV 22:04:30 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 167.536 uVs 22:04:33 I2Y Average and RMS of Measurements VT = -164.813 +- 5.337 uVs I2Y = -15.163 +- 0.491 uTm^2 22:04:37 Wire Position: X0 = 0.0000 mm, Y0 = 0.5000 mm 22:04:41 Integrated Voltage Measurement... 22:04:50 Voltage Offset Correction: v_off_ini = 2959.591 uV, v_off_fin = 2947.645 uV 22:04:50 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -95.765 uVs 22:04:50 Integrated Voltage Measurement... 22:04:59 Voltage Offset Correction: v_off_ini = 2953.522 uV, v_off_fin = 2958.412 uV 22:04:59 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 81.371 uVs 22:04:59 Integrated Voltage Measurement... 22:05:08 Voltage Offset Correction: v_off_ini = 2954.804 uV, v_off_fin = 2937.195 uV 22:05:08 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -87.958 uVs 22:05:08 Integrated Voltage Measurement... 22:05:17 Voltage Offset Correction: v_off_ini = 2939.745 uV, v_off_fin = 2973.469 uV 22:05:17 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = 97.907 uVs 22:05:17 Integrated Voltage Measurement... 22:05:26 Voltage Offset Correction: v_off_ini = 2940.233 uV, v_off_fin = 2940.524 uV 22:05:26 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -57.990 uVs 22:05:26 Integrated Voltage Measurement... 22:05:35 Voltage Offset Correction: v_off_ini = 2950.778 uV, v_off_fin = 2956.931 uV 22:05:35 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 66.027 uVs 22:05:39 I1X Average and RMS of Measurements VT = -81.170 +- 13.666 uVs I1X = 16.235 +- 2.734 uTm 22:05:42 Wire Position: X0 = 0.0000 mm, Y0 = 0.5000 mm 22:05:46 Integrated Voltage Measurement... 22:05:55 Voltage Offset Correction: v_off_ini = 2962.672 uV, v_off_fin = 2940.776 uV 22:05:55 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -156.713 uVs 22:05:55 Integrated Voltage Measurement... 22:06:04 Voltage Offset Correction: v_off_ini = 2976.182 uV, v_off_fin = 2983.589 uV 22:06:04 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 170.658 uVs 22:06:04 Integrated Voltage Measurement... 22:06:13 Voltage Offset Correction: v_off_ini = 2945.448 uV, v_off_fin = 2932.092 uV 22:06:13 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -158.207 uVs 22:06:13 Integrated Voltage Measurement... 22:06:22 Voltage Offset Correction: v_off_ini = 2960.177 uV, v_off_fin = 2952.034 uV 22:06:22 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 168.629 uVs 22:06:22 Integrated Voltage Measurement... 22:06:31 Voltage Offset Correction: v_off_ini = 2941.002 uV, v_off_fin = 2931.332 uV 22:06:31 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -154.858 uVs 22:06:31 Integrated Voltage Measurement... 22:06:40 Voltage Offset Correction: v_off_ini = 2958.998 uV, v_off_fin = 2946.727 uV 22:06:40 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 177.471 uVs 22:06:44 I1Y Average and RMS of Measurements VT = -164.423 +- 1.236 uVs I1Y = -32.881 +- 0.247 uTm 22:06:47 Wire Position: X0 = 0.0000 mm, Y0 = 0.5000 mm 22:06:49 Integrated Voltage Measurement... 22:06:58 Voltage Offset Correction: v_off_ini = 2936.347 uV, v_off_fin = 2941.610 uV 22:06:58 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -64.558 uVs 22:06:58 Integrated Voltage Measurement... 22:07:07 Voltage Offset Correction: v_off_ini = 2946.970 uV, v_off_fin = 2967.223 uV 22:07:07 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 82.391 uVs 22:07:07 Integrated Voltage Measurement... 22:07:16 Voltage Offset Correction: v_off_ini = 2961.091 uV, v_off_fin = 2956.467 uV 22:07:16 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -90.044 uVs 22:07:16 Integrated Voltage Measurement... 22:07:25 Voltage Offset Correction: v_off_ini = 2970.495 uV, v_off_fin = 2962.121 uV 22:07:25 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 41.979 uVs 22:07:25 Integrated Voltage Measurement... 22:07:34 Voltage Offset Correction: v_off_ini = 2973.123 uV, v_off_fin = 2955.679 uV 22:07:34 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -74.487 uVs 22:07:34 Integrated Voltage Measurement... 22:07:43 Voltage Offset Correction: v_off_ini = 2955.644 uV, v_off_fin = 2970.938 uV 22:07:43 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 75.672 uVs 22:07:46 I2X Average and RMS of Measurements VT = -71.522 +- 3.951 uVs I2X = 6.580 +- 0.364 uTm^2 22:07:48 Wire Position: X0 = 0.0000 mm, Y0 = 0.5000 mm 22:07:51 Integrated Voltage Measurement... 22:08:00 Voltage Offset Correction: v_off_ini = 2985.865 uV, v_off_fin = 2973.419 uV 22:08:00 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -157.346 uVs 22:08:00 Integrated Voltage Measurement... 22:08:09 Voltage Offset Correction: v_off_ini = 2967.083 uV, v_off_fin = 2986.237 uV 22:08:09 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 172.219 uVs 22:08:09 Integrated Voltage Measurement... 22:08:18 Voltage Offset Correction: v_off_ini = 2973.894 uV, v_off_fin = 2939.254 uV 22:08:18 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -169.426 uVs 22:08:18 Integrated Voltage Measurement... 22:08:27 Voltage Offset Correction: v_off_ini = 2947.497 uV, v_off_fin = 2962.351 uV 22:08:27 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 169.155 uVs 22:08:27 Integrated Voltage Measurement... 22:08:36 Voltage Offset Correction: v_off_ini = 2964.905 uV, v_off_fin = 2960.936 uV 22:08:36 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -167.459 uVs 22:08:36 Integrated Voltage Measurement... 22:08:45 Voltage Offset Correction: v_off_ini = 2969.779 uV, v_off_fin = 2965.841 uV 22:08:45 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 175.390 uVs 22:08:47 I2Y Average and RMS of Measurements VT = -168.499 +- 2.769 uVs I2Y = -15.502 +- 0.255 uTm^2 22:08:51 Wire Position: X0 = 0.0000 mm, Y0 = 1.0000 mm 22:08:56 Integrated Voltage Measurement... 22:09:05 Voltage Offset Correction: v_off_ini = 2963.359 uV, v_off_fin = 2947.328 uV 22:09:05 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -83.181 uVs 22:09:05 Integrated Voltage Measurement... 22:09:14 Voltage Offset Correction: v_off_ini = 2939.228 uV, v_off_fin = 2940.677 uV 22:09:14 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 62.593 uVs 22:09:14 Integrated Voltage Measurement... 22:09:23 Voltage Offset Correction: v_off_ini = 2913.251 uV, v_off_fin = 2941.475 uV 22:09:23 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -66.453 uVs 22:09:23 Integrated Voltage Measurement... 22:09:32 Voltage Offset Correction: v_off_ini = 2971.335 uV, v_off_fin = 2963.202 uV 22:09:32 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 67.003 uVs 22:09:32 Integrated Voltage Measurement... 22:09:41 Voltage Offset Correction: v_off_ini = 2942.255 uV, v_off_fin = 2956.934 uV 22:09:41 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -74.635 uVs 22:09:41 Integrated Voltage Measurement... 22:09:50 Voltage Offset Correction: v_off_ini = 2955.139 uV, v_off_fin = 2975.799 uV 22:09:50 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 76.245 uVs 22:09:54 I1X Average and RMS of Measurements VT = -71.685 +- 3.657 uVs I1X = 14.337 +- 0.731 uTm 22:09:56 Wire Position: X0 = 0.0000 mm, Y0 = 1.0000 mm 22:10:01 Integrated Voltage Measurement... 22:10:10 Voltage Offset Correction: v_off_ini = 2938.483 uV, v_off_fin = 2910.901 uV 22:10:10 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -171.322 uVs 22:10:10 Integrated Voltage Measurement... 22:10:19 Voltage Offset Correction: v_off_ini = 2945.013 uV, v_off_fin = 2947.388 uV 22:10:19 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 166.564 uVs 22:10:19 Integrated Voltage Measurement... 22:10:28 Voltage Offset Correction: v_off_ini = 2931.579 uV, v_off_fin = 2928.184 uV 22:10:28 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -148.651 uVs 22:10:28 Integrated Voltage Measurement... 22:10:37 Voltage Offset Correction: v_off_ini = 2930.336 uV, v_off_fin = 2943.310 uV 22:10:37 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 166.172 uVs 22:10:37 Integrated Voltage Measurement... 22:10:46 Voltage Offset Correction: v_off_ini = 2914.782 uV, v_off_fin = 2902.254 uV 22:10:46 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -160.600 uVs 22:10:46 Integrated Voltage Measurement... 22:10:54 Voltage Offset Correction: v_off_ini = 2908.349 uV, v_off_fin = 2953.109 uV 22:10:55 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 174.616 uVs 22:10:59 I1Y Average and RMS of Measurements VT = -164.654 +- 5.150 uVs I1Y = -32.928 +- 1.030 uTm 22:11:01 Wire Position: X0 = 0.0000 mm, Y0 = 1.0000 mm 22:11:04 Integrated Voltage Measurement... 22:11:13 Voltage Offset Correction: v_off_ini = 2916.208 uV, v_off_fin = 2920.543 uV 22:11:13 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -71.494 uVs 22:11:13 Integrated Voltage Measurement... 22:11:22 Voltage Offset Correction: v_off_ini = 2921.222 uV, v_off_fin = 2927.635 uV 22:11:22 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 72.263 uVs 22:11:22 Integrated Voltage Measurement... 22:11:31 Voltage Offset Correction: v_off_ini = 2923.580 uV, v_off_fin = 2901.843 uV 22:11:31 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -77.512 uVs 22:11:31 Integrated Voltage Measurement... 22:11:40 Voltage Offset Correction: v_off_ini = 2906.884 uV, v_off_fin = 2911.994 uV 22:11:40 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 47.024 uVs 22:11:40 Integrated Voltage Measurement... 22:11:49 Voltage Offset Correction: v_off_ini = 2907.043 uV, v_off_fin = 2899.596 uV 22:11:49 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -55.651 uVs 22:11:49 Integrated Voltage Measurement... 22:11:58 Voltage Offset Correction: v_off_ini = 2909.403 uV, v_off_fin = 2928.924 uV 22:11:58 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 48.594 uVs 22:12:00 I2X Average and RMS of Measurements VT = -62.090 +- 8.066 uVs I2X = 5.712 +- 0.742 uTm^2 22:12:03 Wire Position: X0 = 0.0000 mm, Y0 = 1.0000 mm 22:12:05 Integrated Voltage Measurement... 22:12:14 Voltage Offset Correction: v_off_ini = 2907.749 uV, v_off_fin = 2916.323 uV 22:12:14 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -166.981 uVs 22:12:14 Integrated Voltage Measurement... 22:12:23 Voltage Offset Correction: v_off_ini = 2913.648 uV, v_off_fin = 2923.240 uV 22:12:23 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 155.515 uVs 22:12:23 Integrated Voltage Measurement... 22:12:32 Voltage Offset Correction: v_off_ini = 2916.403 uV, v_off_fin = 2917.249 uV 22:12:32 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -167.536 uVs 22:12:33 Integrated Voltage Measurement... 22:12:41 Voltage Offset Correction: v_off_ini = 2907.502 uV, v_off_fin = 2918.109 uV 22:12:42 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 177.854 uVs 22:12:42 Integrated Voltage Measurement... 22:12:51 Voltage Offset Correction: v_off_ini = 2931.528 uV, v_off_fin = 2930.225 uV 22:12:51 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -179.188 uVs 22:12:51 Integrated Voltage Measurement... 22:13:00 Voltage Offset Correction: v_off_ini = 2903.780 uV, v_off_fin = 2912.501 uV 22:13:00 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 156.879 uVs 22:13:02 I2Y Average and RMS of Measurements VT = -167.326 +- 4.700 uVs I2Y = -15.394 +- 0.432 uTm^2 22:13:07 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm