SLAC Magnetic Measurements Date: 07-13-2026 Time: 13:30:13 Project: LCLS-II-HE Magnet Type: Phase Shifter Magnet Name: HE_SXPS_15081 Dataset Number: 5 Run Purpose: Tuning Run Type: Run Number: 3 Measurement Device: Stretched Wire Operator: YL Comment: On-Axis 13:30:21 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 13:30:23 Motion Parameters, X Controller, Axis 1 acc = 0.001000 m/s/s dcl = 0.001000 m/s/s vel = 0.001000 m/s pos = 0.000000 m kp = 188.600000 ki = 820.000000 kd = 491.400000 db = 0.000000 13:30:25 Motion Parameters, X Controller, Axis 2 acc = 0.001000 m/s/s dcl = 0.001000 m/s/s vel = 0.001000 m/s pos = 0.000000 m kp = 188.600000 ki = 820.000000 kd = 491.400000 db = 0.000000 13:30:26 Motion Parameters, Y Controller, Axis 1 acc = 0.001000 m/s/s dcl = 0.001000 m/s/s vel = 0.001000 m/s pos = 0.000000 m kp = 188.600000 ki = 820.000000 kd = 491.400000 db = 0.000000 13:30:28 Motion Parameters, Y Controller, Axis 2 acc = 0.001000 m/s/s dcl = 0.001000 m/s/s vel = 0.001000 m/s pos = 0.000000 m kp = 188.600000 ki = 820.000000 kd = 491.400000 db = 0.000000 13:31:35 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 13:31:39 Integrated Voltage Measurement... 13:31:48 Voltage Offset Correction: v_off_ini = 1665.372 uV, v_off_fin = 1645.087 uV 13:31:48 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -127.370 uVs 13:31:48 Integrated Voltage Measurement... 13:31:57 Voltage Offset Correction: v_off_ini = 1635.237 uV, v_off_fin = 1643.796 uV 13:31:57 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 88.855 uVs 13:31:57 Integrated Voltage Measurement... 13:32:06 Voltage Offset Correction: v_off_ini = 1644.665 uV, v_off_fin = 1631.371 uV 13:32:06 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -74.222 uVs 13:32:06 Integrated Voltage Measurement... 13:32:15 Voltage Offset Correction: v_off_ini = 1635.677 uV, v_off_fin = 1654.645 uV 13:32:15 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 84.057 uVs 13:32:15 Integrated Voltage Measurement... 13:32:24 Voltage Offset Correction: v_off_ini = 1631.136 uV, v_off_fin = 1625.100 uV 13:32:24 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -94.968 uVs 13:32:24 Integrated Voltage Measurement... 13:32:33 Voltage Offset Correction: v_off_ini = 1628.661 uV, v_off_fin = 1633.512 uV 13:32:33 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 35.645 uVs 13:32:37 I1X Average and RMS of Measurements VT = -84.186 +- 17.836 uVs I1X = 16.837 +- 3.567 uTm 13:32:40 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 13:32:44 Integrated Voltage Measurement... 13:32:53 Voltage Offset Correction: v_off_ini = 1602.796 uV, v_off_fin = 1597.007 uV 13:32:53 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -291.412 uVs 13:32:53 Integrated Voltage Measurement... 13:33:02 Voltage Offset Correction: v_off_ini = 1622.641 uV, v_off_fin = 1614.932 uV 13:33:02 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 306.136 uVs 13:33:02 Integrated Voltage Measurement... 13:33:11 Voltage Offset Correction: v_off_ini = 1605.581 uV, v_off_fin = 1598.059 uV 13:33:11 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -302.853 uVs 13:33:11 Integrated Voltage Measurement... 13:33:20 Voltage Offset Correction: v_off_ini = 1621.233 uV, v_off_fin = 1631.226 uV 13:33:20 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 287.950 uVs 13:33:20 Integrated Voltage Measurement... 13:33:29 Voltage Offset Correction: v_off_ini = 1629.308 uV, v_off_fin = 1630.814 uV 13:33:29 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -283.070 uVs 13:33:29 Integrated Voltage Measurement... 13:33:38 Voltage Offset Correction: v_off_ini = 1635.127 uV, v_off_fin = 1637.752 uV 13:33:38 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 326.820 uVs 13:33:42 I1Y Average and RMS of Measurements VT = -299.707 +- 3.952 uVs I1Y = -59.942 +- 0.792 uTm 13:33:47 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 13:33:50 Integrated Voltage Measurement... 13:33:59 Voltage Offset Correction: v_off_ini = 1617.601 uV, v_off_fin = 1644.798 uV 13:33:59 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -57.439 uVs 13:33:59 Integrated Voltage Measurement... 13:34:08 Voltage Offset Correction: v_off_ini = 1623.551 uV, v_off_fin = 1614.130 uV 13:34:08 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 53.603 uVs 13:34:08 Integrated Voltage Measurement... 13:34:17 Voltage Offset Correction: v_off_ini = 1616.174 uV, v_off_fin = 1592.822 uV 13:34:17 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -39.911 uVs 13:34:17 Integrated Voltage Measurement... 13:34:26 Voltage Offset Correction: v_off_ini = 1615.505 uV, v_off_fin = 1634.720 uV 13:34:26 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 36.890 uVs 13:34:26 Integrated Voltage Measurement... 13:34:35 Voltage Offset Correction: v_off_ini = 1604.156 uV, v_off_fin = 1600.469 uV 13:34:35 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -54.979 uVs 13:34:35 Integrated Voltage Measurement... 13:34:44 Voltage Offset Correction: v_off_ini = 1603.664 uV, v_off_fin = 1601.813 uV 13:34:44 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 43.247 uVs 13:34:46 I2X Average and RMS of Measurements VT = -47.678 +- 7.063 uVs I2X = 4.386 +- 0.650 uTm^2 13:34:51 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 13:34:54 Integrated Voltage Measurement... 13:35:03 Voltage Offset Correction: v_off_ini = 1601.942 uV, v_off_fin = 1618.538 uV 13:35:03 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -293.370 uVs 13:35:03 Integrated Voltage Measurement... 13:35:12 Voltage Offset Correction: v_off_ini = 1591.517 uV, v_off_fin = 1593.880 uV 13:35:12 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 293.583 uVs 13:35:12 Integrated Voltage Measurement... 13:35:21 Voltage Offset Correction: v_off_ini = 1590.861 uV, v_off_fin = 1601.646 uV 13:35:21 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -288.517 uVs 13:35:21 Integrated Voltage Measurement... 13:35:30 Voltage Offset Correction: v_off_ini = 1633.960 uV, v_off_fin = 1609.342 uV 13:35:30 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 305.386 uVs 13:35:30 Integrated Voltage Measurement... 13:35:39 Voltage Offset Correction: v_off_ini = 1605.991 uV, v_off_fin = 1575.008 uV 13:35:39 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -299.553 uVs 13:35:39 Integrated Voltage Measurement... 13:35:48 Voltage Offset Correction: v_off_ini = 1621.514 uV, v_off_fin = 1603.930 uV 13:35:48 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 314.848 uVs 13:35:51 I2Y Average and RMS of Measurements VT = -299.210 +- 5.826 uVs I2Y = -27.527 +- 0.536 uTm^2 13:36:01 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 13:36:06 Integrated Voltage Measurement... 13:36:15 Voltage Offset Correction: v_off_ini = 1622.191 uV, v_off_fin = 1634.823 uV 13:36:15 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -75.197 uVs 13:36:15 Integrated Voltage Measurement... 13:36:24 Voltage Offset Correction: v_off_ini = 1619.347 uV, v_off_fin = 1645.283 uV 13:36:24 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 65.822 uVs 13:36:24 Integrated Voltage Measurement... 13:36:33 Voltage Offset Correction: v_off_ini = 1629.030 uV, v_off_fin = 1643.702 uV 13:36:33 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -87.356 uVs 13:36:33 Integrated Voltage Measurement... 13:36:42 Voltage Offset Correction: v_off_ini = 1639.866 uV, v_off_fin = 1633.391 uV 13:36:42 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 53.719 uVs 13:36:42 Integrated Voltage Measurement... 13:36:51 Voltage Offset Correction: v_off_ini = 1646.929 uV, v_off_fin = 1648.026 uV 13:36:51 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -64.124 uVs 13:36:51 Integrated Voltage Measurement... 13:37:00 Voltage Offset Correction: v_off_ini = 1630.861 uV, v_off_fin = 1638.758 uV 13:37:00 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 78.566 uVs 13:37:04 I1X Average and RMS of Measurements VT = -70.797 +- 0.387 uVs I1X = 14.160 +- 0.077 uTm 13:37:06 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 13:37:11 Integrated Voltage Measurement... 13:37:19 Voltage Offset Correction: v_off_ini = 1654.740 uV, v_off_fin = 1638.695 uV 13:37:19 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -299.564 uVs 13:37:19 Integrated Voltage Measurement... 13:37:28 Voltage Offset Correction: v_off_ini = 1628.492 uV, v_off_fin = 1645.659 uV 13:37:28 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 286.782 uVs 13:37:28 Integrated Voltage Measurement... 13:37:37 Voltage Offset Correction: v_off_ini = 1634.102 uV, v_off_fin = 1629.205 uV 13:37:37 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -291.473 uVs 13:37:37 Integrated Voltage Measurement... 13:37:46 Voltage Offset Correction: v_off_ini = 1605.545 uV, v_off_fin = 1618.120 uV 13:37:46 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 300.600 uVs 13:37:46 Integrated Voltage Measurement... 13:37:55 Voltage Offset Correction: v_off_ini = 1604.822 uV, v_off_fin = 1600.835 uV 13:37:55 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -304.277 uVs 13:37:55 Integrated Voltage Measurement... 13:38:04 Voltage Offset Correction: v_off_ini = 1620.224 uV, v_off_fin = 1628.438 uV 13:38:04 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 284.452 uVs 13:38:09 I1Y Average and RMS of Measurements VT = -294.525 +- 1.174 uVs I1Y = -58.905 +- 0.235 uTm 13:38:14 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 13:38:16 Integrated Voltage Measurement... 13:38:25 Voltage Offset Correction: v_off_ini = 1617.837 uV, v_off_fin = 1595.007 uV 13:38:25 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -40.136 uVs 13:38:25 Integrated Voltage Measurement... 13:38:34 Voltage Offset Correction: v_off_ini = 1606.357 uV, v_off_fin = 1626.548 uV 13:38:34 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 55.771 uVs 13:38:34 Integrated Voltage Measurement... 13:38:43 Voltage Offset Correction: v_off_ini = 1602.103 uV, v_off_fin = 1580.712 uV 13:38:43 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -50.605 uVs 13:38:43 Integrated Voltage Measurement... 13:38:52 Voltage Offset Correction: v_off_ini = 1587.734 uV, v_off_fin = 1616.788 uV 13:38:52 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 37.563 uVs 13:38:52 Integrated Voltage Measurement... 13:39:01 Voltage Offset Correction: v_off_ini = 1596.359 uV, v_off_fin = 1599.544 uV 13:39:01 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -36.592 uVs 13:39:01 Integrated Voltage Measurement... 13:39:10 Voltage Offset Correction: v_off_ini = 1605.797 uV, v_off_fin = 1594.216 uV 13:39:10 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 22.418 uVs 13:39:13 I2X Average and RMS of Measurements VT = -40.514 +- 7.943 uVs I2X = 3.727 +- 0.731 uTm^2 13:39:18 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 13:39:20 Integrated Voltage Measurement... 13:39:29 Voltage Offset Correction: v_off_ini = 1592.475 uV, v_off_fin = 1596.641 uV 13:39:29 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -310.860 uVs 13:39:29 Integrated Voltage Measurement... 13:39:38 Voltage Offset Correction: v_off_ini = 1590.884 uV, v_off_fin = 1618.952 uV 13:39:38 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 290.666 uVs 13:39:38 Integrated Voltage Measurement... 13:39:47 Voltage Offset Correction: v_off_ini = 1600.622 uV, v_off_fin = 1576.786 uV 13:39:47 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -306.286 uVs 13:39:48 Integrated Voltage Measurement... 13:39:56 Voltage Offset Correction: v_off_ini = 1586.613 uV, v_off_fin = 1593.194 uV 13:39:56 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 294.585 uVs 13:39:57 Integrated Voltage Measurement... 13:40:06 Voltage Offset Correction: v_off_ini = 1581.748 uV, v_off_fin = 1570.276 uV 13:40:06 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -295.406 uVs 13:40:06 Integrated Voltage Measurement... 13:40:15 Voltage Offset Correction: v_off_ini = 1592.830 uV, v_off_fin = 1610.112 uV 13:40:15 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 300.739 uVs 13:40:17 I2Y Average and RMS of Measurements VT = -299.757 +- 1.199 uVs I2Y = -27.578 +- 0.110 uTm^2 13:40:28 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 13:40:32 Integrated Voltage Measurement... 13:40:41 Voltage Offset Correction: v_off_ini = 1608.086 uV, v_off_fin = 1619.324 uV 13:40:41 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.4999 mm dXY = -0.4999 mm VT = -85.512 uVs 13:40:41 Integrated Voltage Measurement... 13:40:50 Voltage Offset Correction: v_off_ini = 1604.865 uV, v_off_fin = 1608.376 uV 13:40:50 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = 71.090 uVs 13:40:50 Integrated Voltage Measurement... 13:40:59 Voltage Offset Correction: v_off_ini = 1628.887 uV, v_off_fin = 1633.734 uV 13:40:59 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -69.142 uVs 13:40:59 Integrated Voltage Measurement... 13:41:08 Voltage Offset Correction: v_off_ini = 1631.923 uV, v_off_fin = 1656.503 uV 13:41:08 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = 33.194 uVs 13:41:08 Integrated Voltage Measurement... 13:41:17 Voltage Offset Correction: v_off_ini = 1646.287 uV, v_off_fin = 1637.803 uV 13:41:17 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -54.195 uVs 13:41:17 Integrated Voltage Measurement... 13:41:26 Voltage Offset Correction: v_off_ini = 1635.652 uV, v_off_fin = 1644.995 uV 13:41:26 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 69.189 uVs 13:41:30 I1X Average and RMS of Measurements VT = -63.720 +- 11.169 uVs I1X = 12.745 +- 2.235 uTm 13:41:33 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 13:41:37 Integrated Voltage Measurement... 13:41:46 Voltage Offset Correction: v_off_ini = 1636.995 uV, v_off_fin = 1607.063 uV 13:41:46 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -302.041 uVs 13:41:46 Integrated Voltage Measurement... 13:41:55 Voltage Offset Correction: v_off_ini = 1634.995 uV, v_off_fin = 1643.888 uV 13:41:55 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 304.546 uVs 13:41:55 Integrated Voltage Measurement... 13:42:04 Voltage Offset Correction: v_off_ini = 1594.859 uV, v_off_fin = 1591.443 uV 13:42:04 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -297.707 uVs 13:42:04 Integrated Voltage Measurement... 13:42:13 Voltage Offset Correction: v_off_ini = 1621.201 uV, v_off_fin = 1630.521 uV 13:42:13 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 303.267 uVs 13:42:13 Integrated Voltage Measurement... 13:42:22 Voltage Offset Correction: v_off_ini = 1591.891 uV, v_off_fin = 1573.515 uV 13:42:22 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -291.534 uVs 13:42:22 Integrated Voltage Measurement... 13:42:31 Voltage Offset Correction: v_off_ini = 1580.561 uV, v_off_fin = 1596.598 uV 13:42:31 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 285.798 uVs 13:42:35 I1Y Average and RMS of Measurements VT = -297.482 +- 6.339 uVs I1Y = -59.497 +- 1.266 uTm 13:42:40 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 13:42:43 Integrated Voltage Measurement... 13:42:52 Voltage Offset Correction: v_off_ini = 1574.720 uV, v_off_fin = 1591.893 uV 13:42:52 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -21.920 uVs 13:42:52 Integrated Voltage Measurement... 13:43:01 Voltage Offset Correction: v_off_ini = 1588.959 uV, v_off_fin = 1620.496 uV 13:43:01 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 41.282 uVs 13:43:01 Integrated Voltage Measurement... 13:43:10 Voltage Offset Correction: v_off_ini = 1599.744 uV, v_off_fin = 1610.961 uV 13:43:10 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -53.515 uVs 13:43:10 Integrated Voltage Measurement... 13:43:19 Voltage Offset Correction: v_off_ini = 1614.418 uV, v_off_fin = 1611.596 uV 13:43:19 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 22.902 uVs 13:43:19 Integrated Voltage Measurement... 13:43:28 Voltage Offset Correction: v_off_ini = 1589.194 uV, v_off_fin = 1582.166 uV 13:43:28 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -48.336 uVs 13:43:28 Integrated Voltage Measurement... 13:43:37 Voltage Offset Correction: v_off_ini = 1598.544 uV, v_off_fin = 1610.541 uV 13:43:37 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 35.179 uVs 13:43:39 I2X Average and RMS of Measurements VT = -37.189 +- 4.208 uVs I2X = 3.421 +- 0.387 uTm^2 13:43:44 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 13:43:47 Integrated Voltage Measurement... 13:43:56 Voltage Offset Correction: v_off_ini = 1604.158 uV, v_off_fin = 1609.457 uV 13:43:56 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -323.674 uVs 13:43:56 Integrated Voltage Measurement... 13:44:05 Voltage Offset Correction: v_off_ini = 1587.162 uV, v_off_fin = 1593.848 uV 13:44:05 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 301.432 uVs 13:44:05 Integrated Voltage Measurement... 13:44:14 Voltage Offset Correction: v_off_ini = 1586.240 uV, v_off_fin = 1587.686 uV 13:44:14 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = -277.558 uVs 13:44:14 Integrated Voltage Measurement... 13:44:23 Voltage Offset Correction: v_off_ini = 1595.659 uV, v_off_fin = 1598.763 uV 13:44:23 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 295.500 uVs 13:44:23 Integrated Voltage Measurement... 13:44:32 Voltage Offset Correction: v_off_ini = 1612.092 uV, v_off_fin = 1608.664 uV 13:44:32 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -316.066 uVs 13:44:32 Integrated Voltage Measurement... 13:44:41 Voltage Offset Correction: v_off_ini = 1583.174 uV, v_off_fin = 1586.853 uV 13:44:41 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 292.287 uVs 13:44:43 I2Y Average and RMS of Measurements VT = -301.086 +- 10.846 uVs I2Y = -27.700 +- 0.997 uTm^2 13:44:54 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 13:44:58 Integrated Voltage Measurement... 13:45:07 Voltage Offset Correction: v_off_ini = 1597.053 uV, v_off_fin = 1579.906 uV 13:45:07 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.4998 mm dXY = -0.4999 mm VT = -70.429 uVs 13:45:07 Integrated Voltage Measurement... 13:45:16 Voltage Offset Correction: v_off_ini = 1569.402 uV, v_off_fin = 1557.698 uV 13:45:16 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = 67.354 uVs 13:45:16 Integrated Voltage Measurement... 13:45:25 Voltage Offset Correction: v_off_ini = 1558.388 uV, v_off_fin = 1572.600 uV 13:45:25 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -61.193 uVs 13:45:25 Integrated Voltage Measurement... 13:45:34 Voltage Offset Correction: v_off_ini = 1586.290 uV, v_off_fin = 1584.764 uV 13:45:34 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 66.043 uVs 13:45:34 Integrated Voltage Measurement... 13:45:43 Voltage Offset Correction: v_off_ini = 1581.769 uV, v_off_fin = 1580.013 uV 13:45:43 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -50.337 uVs 13:45:43 Integrated Voltage Measurement... 13:45:52 Voltage Offset Correction: v_off_ini = 1566.117 uV, v_off_fin = 1578.001 uV 13:45:52 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 60.613 uVs 13:45:57 I1X Average and RMS of Measurements VT = -62.662 +- 5.519 uVs I1X = 12.534 +- 1.105 uTm 13:45:59 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 13:46:03 Integrated Voltage Measurement... 13:46:12 Voltage Offset Correction: v_off_ini = 1572.289 uV, v_off_fin = 1557.462 uV 13:46:12 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -279.765 uVs 13:46:12 Integrated Voltage Measurement... 13:46:21 Voltage Offset Correction: v_off_ini = 1569.243 uV, v_off_fin = 1593.013 uV 13:46:21 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 276.765 uVs 13:46:21 Integrated Voltage Measurement... 13:46:30 Voltage Offset Correction: v_off_ini = 1558.407 uV, v_off_fin = 1541.337 uV 13:46:30 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -298.180 uVs 13:46:30 Integrated Voltage Measurement... 13:46:39 Voltage Offset Correction: v_off_ini = 1575.382 uV, v_off_fin = 1590.289 uV 13:46:39 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 259.573 uVs 13:46:39 Integrated Voltage Measurement... 13:46:48 Voltage Offset Correction: v_off_ini = 1603.615 uV, v_off_fin = 1573.406 uV 13:46:48 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -279.858 uVs 13:46:49 Integrated Voltage Measurement... 13:46:58 Voltage Offset Correction: v_off_ini = 1586.004 uV, v_off_fin = 1595.042 uV 13:46:58 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 282.887 uVs 13:47:02 I1Y Average and RMS of Measurements VT = -279.505 +- 1.344 uVs I1Y = -55.900 +- 0.268 uTm 13:47:07 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 13:47:09 Integrated Voltage Measurement... 13:47:18 Voltage Offset Correction: v_off_ini = 1582.211 uV, v_off_fin = 1556.347 uV 13:47:18 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -30.504 uVs 13:47:18 Integrated Voltage Measurement... 13:47:27 Voltage Offset Correction: v_off_ini = 1599.082 uV, v_off_fin = 1597.686 uV 13:47:27 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 35.694 uVs 13:47:28 Integrated Voltage Measurement... 13:47:36 Voltage Offset Correction: v_off_ini = 1575.844 uV, v_off_fin = 1578.238 uV 13:47:37 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -21.667 uVs 13:47:37 Integrated Voltage Measurement... 13:47:46 Voltage Offset Correction: v_off_ini = 1574.339 uV, v_off_fin = 1585.587 uV 13:47:46 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 46.352 uVs 13:47:46 Integrated Voltage Measurement... 13:47:55 Voltage Offset Correction: v_off_ini = 1572.826 uV, v_off_fin = 1569.371 uV 13:47:55 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -32.143 uVs 13:47:55 Integrated Voltage Measurement... 13:48:04 Voltage Offset Correction: v_off_ini = 1588.672 uV, v_off_fin = 1592.843 uV 13:48:04 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 39.823 uVs 13:48:06 I2X Average and RMS of Measurements VT = -34.364 +- 1.204 uVs I2X = 3.161 +- 0.111 uTm^2 13:48:11 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 13:48:14 Integrated Voltage Measurement... 13:48:23 Voltage Offset Correction: v_off_ini = 1553.689 uV, v_off_fin = 1547.851 uV 13:48:23 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -331.310 uVs 13:48:23 Integrated Voltage Measurement... 13:48:32 Voltage Offset Correction: v_off_ini = 1565.578 uV, v_off_fin = 1562.930 uV 13:48:32 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 260.813 uVs 13:48:32 Integrated Voltage Measurement... 13:48:41 Voltage Offset Correction: v_off_ini = 1558.734 uV, v_off_fin = 1548.265 uV 13:48:41 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -291.805 uVs 13:48:41 Integrated Voltage Measurement... 13:48:50 Voltage Offset Correction: v_off_ini = 1542.586 uV, v_off_fin = 1564.297 uV 13:48:50 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 278.725 uVs 13:48:50 Integrated Voltage Measurement... 13:48:59 Voltage Offset Correction: v_off_ini = 1551.803 uV, v_off_fin = 1562.304 uV 13:48:59 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -284.967 uVs 13:48:59 Integrated Voltage Measurement... 13:49:08 Voltage Offset Correction: v_off_ini = 1530.740 uV, v_off_fin = 1533.548 uV 13:49:08 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 302.324 uVs 13:49:10 I2Y Average and RMS of Measurements VT = -291.657 +- 4.626 uVs I2Y = -26.832 +- 0.426 uTm^2 13:49:21 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 13:49:25 Integrated Voltage Measurement... 13:49:34 Voltage Offset Correction: v_off_ini = 1557.928 uV, v_off_fin = 1550.360 uV 13:49:34 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -50.013 uVs 13:49:34 Integrated Voltage Measurement... 13:49:43 Voltage Offset Correction: v_off_ini = 1538.633 uV, v_off_fin = 1536.891 uV 13:49:43 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 63.794 uVs 13:49:43 Integrated Voltage Measurement... 13:49:52 Voltage Offset Correction: v_off_ini = 1529.967 uV, v_off_fin = 1526.964 uV 13:49:52 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -85.109 uVs 13:49:52 Integrated Voltage Measurement... 13:50:01 Voltage Offset Correction: v_off_ini = 1537.981 uV, v_off_fin = 1517.655 uV 13:50:01 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 34.225 uVs 13:50:01 Integrated Voltage Measurement... 13:50:10 Voltage Offset Correction: v_off_ini = 1526.445 uV, v_off_fin = 1522.371 uV 13:50:10 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -52.532 uVs 13:50:10 Integrated Voltage Measurement... 13:50:19 Voltage Offset Correction: v_off_ini = 1519.792 uV, v_off_fin = 1502.524 uV 13:50:19 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 85.199 uVs 13:50:23 I1X Average and RMS of Measurements VT = -61.812 +- 5.114 uVs I1X = 12.363 +- 1.022 uTm 13:50:26 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 13:50:30 Integrated Voltage Measurement... 13:50:39 Voltage Offset Correction: v_off_ini = 1496.864 uV, v_off_fin = 1500.173 uV 13:50:39 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -267.243 uVs 13:50:39 Integrated Voltage Measurement... 13:50:48 Voltage Offset Correction: v_off_ini = 1494.170 uV, v_off_fin = 1522.568 uV 13:50:48 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 278.601 uVs 13:50:48 Integrated Voltage Measurement... 13:50:57 Voltage Offset Correction: v_off_ini = 1535.182 uV, v_off_fin = 1523.869 uV 13:50:57 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -264.646 uVs 13:50:57 Integrated Voltage Measurement... 13:51:06 Voltage Offset Correction: v_off_ini = 1523.184 uV, v_off_fin = 1513.981 uV 13:51:06 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 291.828 uVs 13:51:06 Integrated Voltage Measurement... 13:51:15 Voltage Offset Correction: v_off_ini = 1503.255 uV, v_off_fin = 1497.495 uV 13:51:15 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -271.156 uVs 13:51:15 Integrated Voltage Measurement... 13:51:24 Voltage Offset Correction: v_off_ini = 1516.485 uV, v_off_fin = 1542.649 uV 13:51:24 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 268.819 uVs 13:51:28 I1Y Average and RMS of Measurements VT = -273.715 +- 3.414 uVs I1Y = -54.738 +- 0.684 uTm 13:51:34 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 13:51:36 Integrated Voltage Measurement... 13:51:45 Voltage Offset Correction: v_off_ini = 1536.834 uV, v_off_fin = 1512.814 uV 13:51:45 Integrated Voltage Measurement dXYus = -1.0002 mm, dXYds = 0.0000 mm dXY = -1.0002 mm VT = -18.267 uVs 13:51:45 Integrated Voltage Measurement... 13:51:54 Voltage Offset Correction: v_off_ini = 1536.022 uV, v_off_fin = 1525.338 uV 13:51:54 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 23.702 uVs 13:51:54 Integrated Voltage Measurement... 13:52:03 Voltage Offset Correction: v_off_ini = 1522.540 uV, v_off_fin = 1506.851 uV 13:52:03 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -35.162 uVs 13:52:03 Integrated Voltage Measurement... 13:52:12 Voltage Offset Correction: v_off_ini = 1503.501 uV, v_off_fin = 1507.343 uV 13:52:12 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 44.517 uVs 13:52:12 Integrated Voltage Measurement... 13:52:21 Voltage Offset Correction: v_off_ini = 1525.279 uV, v_off_fin = 1509.459 uV 13:52:21 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -38.065 uVs 13:52:21 Integrated Voltage Measurement... 13:52:30 Voltage Offset Correction: v_off_ini = 1521.999 uV, v_off_fin = 1556.434 uV 13:52:30 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 18.951 uVs 13:52:33 I2X Average and RMS of Measurements VT = -29.777 +- 7.750 uVs I2X = 2.739 +- 0.713 uTm^2 13:52:38 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 13:52:40 Integrated Voltage Measurement... 13:52:49 Voltage Offset Correction: v_off_ini = 1512.129 uV, v_off_fin = 1516.234 uV 13:52:49 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -294.746 uVs 13:52:49 Integrated Voltage Measurement... 13:52:58 Voltage Offset Correction: v_off_ini = 1506.376 uV, v_off_fin = 1507.021 uV 13:52:58 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 269.663 uVs 13:52:58 Integrated Voltage Measurement... 13:53:07 Voltage Offset Correction: v_off_ini = 1481.425 uV, v_off_fin = 1492.758 uV 13:53:07 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -245.909 uVs 13:53:07 Integrated Voltage Measurement... 13:53:16 Voltage Offset Correction: v_off_ini = 1492.136 uV, v_off_fin = 1502.713 uV 13:53:16 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 278.194 uVs 13:53:16 Integrated Voltage Measurement... 13:53:25 Voltage Offset Correction: v_off_ini = 1502.035 uV, v_off_fin = 1524.820 uV 13:53:25 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -255.937 uVs 13:53:25 Integrated Voltage Measurement... 13:53:34 Voltage Offset Correction: v_off_ini = 1518.315 uV, v_off_fin = 1516.660 uV 13:53:35 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 245.003 uVs 13:53:37 I2Y Average and RMS of Measurements VT = -264.908 +- 13.112 uVs I2Y = -24.372 +- 1.206 uTm^2 13:53:48 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 13:53:52 Integrated Voltage Measurement... 13:54:01 Voltage Offset Correction: v_off_ini = 1514.020 uV, v_off_fin = 1515.541 uV 13:54:01 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -44.437 uVs 13:54:01 Integrated Voltage Measurement... 13:54:10 Voltage Offset Correction: v_off_ini = 1516.561 uV, v_off_fin = 1525.199 uV 13:54:10 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -10.033 uVs 13:54:10 Integrated Voltage Measurement... 13:54:19 Voltage Offset Correction: v_off_ini = 1511.979 uV, v_off_fin = 1496.003 uV 13:54:19 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -68.235 uVs 13:54:19 Integrated Voltage Measurement... 13:54:28 Voltage Offset Correction: v_off_ini = 1502.146 uV, v_off_fin = 1501.188 uV 13:54:28 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 65.285 uVs 13:54:28 Integrated Voltage Measurement... 13:54:37 Voltage Offset Correction: v_off_ini = 1490.089 uV, v_off_fin = 1477.768 uV 13:54:37 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -56.351 uVs 13:54:37 Integrated Voltage Measurement... 13:54:46 Voltage Offset Correction: v_off_ini = 1501.434 uV, v_off_fin = 1518.972 uV 13:54:46 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 81.191 uVs 13:54:50 I1X Average and RMS of Measurements VT = -50.911 +- 23.850 uVs I1X = 10.182 +- 4.770 uTm 13:54:53 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 13:54:57 Integrated Voltage Measurement... 13:55:06 Voltage Offset Correction: v_off_ini = 1493.639 uV, v_off_fin = 1491.136 uV 13:55:06 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -263.073 uVs 13:55:06 Integrated Voltage Measurement... 13:55:15 Voltage Offset Correction: v_off_ini = 1526.390 uV, v_off_fin = 1521.529 uV 13:55:15 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 261.115 uVs 13:55:15 Integrated Voltage Measurement... 13:55:24 Voltage Offset Correction: v_off_ini = 1513.449 uV, v_off_fin = 1484.154 uV 13:55:24 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -263.001 uVs 13:55:24 Integrated Voltage Measurement... 13:55:33 Voltage Offset Correction: v_off_ini = 1481.492 uV, v_off_fin = 1497.753 uV 13:55:33 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 258.632 uVs 13:55:33 Integrated Voltage Measurement... 13:55:42 Voltage Offset Correction: v_off_ini = 1494.248 uV, v_off_fin = 1484.733 uV 13:55:42 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -249.089 uVs 13:55:42 Integrated Voltage Measurement... 13:55:51 Voltage Offset Correction: v_off_ini = 1509.891 uV, v_off_fin = 1516.076 uV 13:55:51 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 274.541 uVs 13:55:55 I1Y Average and RMS of Measurements VT = -261.575 +- 0.548 uVs I1Y = -52.311 +- 0.110 uTm 13:56:00 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 13:56:03 Integrated Voltage Measurement... 13:56:12 Voltage Offset Correction: v_off_ini = 1500.376 uV, v_off_fin = 1485.668 uV 13:56:12 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -7.574 uVs 13:56:12 Integrated Voltage Measurement... 13:56:21 Voltage Offset Correction: v_off_ini = 1496.007 uV, v_off_fin = 1511.497 uV 13:56:21 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -2.143 uVs 13:56:21 Integrated Voltage Measurement... 13:56:30 Voltage Offset Correction: v_off_ini = 1507.963 uV, v_off_fin = 1501.936 uV 13:56:30 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -21.476 uVs 13:56:30 Integrated Voltage Measurement... 13:56:39 Voltage Offset Correction: v_off_ini = 1497.018 uV, v_off_fin = 1509.051 uV 13:56:39 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 4.796 uVs 13:56:39 Integrated Voltage Measurement... 13:56:48 Voltage Offset Correction: v_off_ini = 1503.169 uV, v_off_fin = 1479.400 uV 13:56:48 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -32.107 uVs 13:56:48 Integrated Voltage Measurement... 13:56:57 Voltage Offset Correction: v_off_ini = 1522.078 uV, v_off_fin = 1523.846 uV 13:56:57 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 34.568 uVs 13:56:59 I2X Average and RMS of Measurements VT = -16.396 +- 12.712 uVs I2X = 1.508 +- 1.170 uTm^2 13:57:04 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 13:57:07 Integrated Voltage Measurement... 13:57:16 Voltage Offset Correction: v_off_ini = 1503.365 uV, v_off_fin = 1494.905 uV 13:57:16 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = -261.039 uVs 13:57:16 Integrated Voltage Measurement... 13:57:25 Voltage Offset Correction: v_off_ini = 1498.133 uV, v_off_fin = 1507.343 uV 13:57:25 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 267.519 uVs 13:57:25 Integrated Voltage Measurement... 13:57:34 Voltage Offset Correction: v_off_ini = 1508.835 uV, v_off_fin = 1486.483 uV 13:57:34 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -289.071 uVs 13:57:34 Integrated Voltage Measurement... 13:57:43 Voltage Offset Correction: v_off_ini = 1479.438 uV, v_off_fin = 1482.595 uV 13:57:43 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 260.829 uVs 13:57:43 Integrated Voltage Measurement... 13:57:52 Voltage Offset Correction: v_off_ini = 1497.945 uV, v_off_fin = 1520.359 uV 13:57:52 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -265.521 uVs 13:57:52 Integrated Voltage Measurement... 13:58:01 Voltage Offset Correction: v_off_ini = 1526.444 uV, v_off_fin = 1517.155 uV 13:58:01 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = 296.719 uVs 13:58:03 I2Y Average and RMS of Measurements VT = -273.450 +- 6.956 uVs I2Y = -25.158 +- 0.640 uTm^2 13:58:14 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 13:58:18 Integrated Voltage Measurement... 13:58:27 Voltage Offset Correction: v_off_ini = 1514.245 uV, v_off_fin = 1502.668 uV 13:58:27 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -63.263 uVs 13:58:27 Integrated Voltage Measurement... 13:58:36 Voltage Offset Correction: v_off_ini = 1488.196 uV, v_off_fin = 1508.091 uV 13:58:36 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 75.467 uVs 13:58:36 Integrated Voltage Measurement... 13:58:45 Voltage Offset Correction: v_off_ini = 1525.611 uV, v_off_fin = 1506.045 uV 13:58:45 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -53.050 uVs 13:58:45 Integrated Voltage Measurement... 13:58:54 Voltage Offset Correction: v_off_ini = 1511.060 uV, v_off_fin = 1500.619 uV 13:58:54 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 54.315 uVs 13:58:54 Integrated Voltage Measurement... 13:59:03 Voltage Offset Correction: v_off_ini = 1503.447 uV, v_off_fin = 1493.493 uV 13:59:03 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -73.831 uVs 13:59:03 Integrated Voltage Measurement... 13:59:12 Voltage Offset Correction: v_off_ini = 1468.265 uV, v_off_fin = 1477.917 uV 13:59:12 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 55.789 uVs 13:59:17 I1X Average and RMS of Measurements VT = -62.619 +- 6.587 uVs I1X = 12.524 +- 1.317 uTm 13:59:19 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 13:59:23 Integrated Voltage Measurement... 13:59:32 Voltage Offset Correction: v_off_ini = 1487.160 uV, v_off_fin = 1472.797 uV 13:59:32 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -270.944 uVs 13:59:32 Integrated Voltage Measurement... 13:59:41 Voltage Offset Correction: v_off_ini = 1478.559 uV, v_off_fin = 1508.807 uV 13:59:41 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 252.626 uVs 13:59:41 Integrated Voltage Measurement... 13:59:50 Voltage Offset Correction: v_off_ini = 1506.783 uV, v_off_fin = 1506.810 uV 13:59:50 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -254.426 uVs 13:59:50 Integrated Voltage Measurement... 13:59:59 Voltage Offset Correction: v_off_ini = 1496.413 uV, v_off_fin = 1499.824 uV 13:59:59 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 285.777 uVs 13:59:59 Integrated Voltage Measurement... 14:00:08 Voltage Offset Correction: v_off_ini = 1491.447 uV, v_off_fin = 1474.675 uV 14:00:08 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -270.449 uVs 14:00:08 Integrated Voltage Measurement... 14:00:17 Voltage Offset Correction: v_off_ini = 1457.977 uV, v_off_fin = 1478.032 uV 14:00:17 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 264.941 uVs 14:00:21 I1Y Average and RMS of Measurements VT = -266.527 +- 3.494 uVs I1Y = -53.303 +- 0.701 uTm 14:00:26 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 14:00:29 Integrated Voltage Measurement... 14:00:38 Voltage Offset Correction: v_off_ini = 1473.843 uV, v_off_fin = 1470.240 uV 14:00:38 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -31.822 uVs 14:00:38 Integrated Voltage Measurement... 14:00:47 Voltage Offset Correction: v_off_ini = 1473.764 uV, v_off_fin = 1488.638 uV 14:00:47 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 5.780 uVs 14:00:47 Integrated Voltage Measurement... 14:00:56 Voltage Offset Correction: v_off_ini = 1490.234 uV, v_off_fin = 1463.884 uV 14:00:56 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -36.772 uVs 14:00:56 Integrated Voltage Measurement... 14:01:05 Voltage Offset Correction: v_off_ini = 1462.936 uV, v_off_fin = 1502.979 uV 14:01:05 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 24.616 uVs 14:01:05 Integrated Voltage Measurement... 14:01:14 Voltage Offset Correction: v_off_ini = 1475.950 uV, v_off_fin = 1477.167 uV 14:01:14 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -1.984 uVs 14:01:14 Integrated Voltage Measurement... 14:01:23 Voltage Offset Correction: v_off_ini = 1487.373 uV, v_off_fin = 1506.916 uV 14:01:23 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 16.306 uVs 14:01:26 I2X Average and RMS of Measurements VT = -19.546 +- 8.813 uVs I2X = 1.798 +- 0.811 uTm^2 14:01:31 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 14:01:33 Integrated Voltage Measurement... 14:01:42 Voltage Offset Correction: v_off_ini = 1490.741 uV, v_off_fin = 1492.151 uV 14:01:42 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -285.447 uVs 14:01:42 Integrated Voltage Measurement... 14:01:51 Voltage Offset Correction: v_off_ini = 1497.044 uV, v_off_fin = 1501.938 uV 14:01:51 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 245.658 uVs 14:01:51 Integrated Voltage Measurement... 14:02:00 Voltage Offset Correction: v_off_ini = 1468.627 uV, v_off_fin = 1475.094 uV 14:02:00 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -263.754 uVs 14:02:00 Integrated Voltage Measurement... 14:02:09 Voltage Offset Correction: v_off_ini = 1478.180 uV, v_off_fin = 1481.496 uV 14:02:09 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = 275.039 uVs 14:02:09 Integrated Voltage Measurement... 14:02:18 Voltage Offset Correction: v_off_ini = 1482.595 uV, v_off_fin = 1472.373 uV 14:02:18 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -242.608 uVs 14:02:18 Integrated Voltage Measurement... 14:02:27 Voltage Offset Correction: v_off_ini = 1466.463 uV, v_off_fin = 1458.452 uV 14:02:27 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 256.022 uVs 14:02:30 I2Y Average and RMS of Measurements VT = -261.421 +- 8.703 uVs I2Y = -24.051 +- 0.801 uTm^2 14:02:41 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 14:02:45 Integrated Voltage Measurement... 14:02:54 Voltage Offset Correction: v_off_ini = 1483.129 uV, v_off_fin = 1463.712 uV 14:02:54 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -72.234 uVs 14:02:54 Integrated Voltage Measurement... 14:03:03 Voltage Offset Correction: v_off_ini = 1463.170 uV, v_off_fin = 1474.373 uV 14:03:03 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 57.000 uVs 14:03:03 Integrated Voltage Measurement... 14:03:12 Voltage Offset Correction: v_off_ini = 1466.461 uV, v_off_fin = 1463.288 uV 14:03:12 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -48.976 uVs 14:03:12 Integrated Voltage Measurement... 14:03:21 Voltage Offset Correction: v_off_ini = 1458.117 uV, v_off_fin = 1484.249 uV 14:03:21 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 60.678 uVs 14:03:21 Integrated Voltage Measurement... 14:03:30 Voltage Offset Correction: v_off_ini = 1464.495 uV, v_off_fin = 1454.121 uV 14:03:30 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -36.926 uVs 14:03:30 Integrated Voltage Measurement... 14:03:39 Voltage Offset Correction: v_off_ini = 1457.944 uV, v_off_fin = 1480.112 uV 14:03:39 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 60.904 uVs 14:03:43 I1X Average and RMS of Measurements VT = -56.120 +- 6.475 uVs I1X = 11.224 +- 1.295 uTm 14:03:45 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 14:03:50 Integrated Voltage Measurement... 14:03:59 Voltage Offset Correction: v_off_ini = 1454.613 uV, v_off_fin = 1440.093 uV 14:03:59 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -254.357 uVs 14:03:59 Integrated Voltage Measurement... 14:04:08 Voltage Offset Correction: v_off_ini = 1442.633 uV, v_off_fin = 1471.369 uV 14:04:08 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 259.144 uVs 14:04:08 Integrated Voltage Measurement... 14:04:17 Voltage Offset Correction: v_off_ini = 1465.179 uV, v_off_fin = 1445.931 uV 14:04:17 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -247.353 uVs 14:04:17 Integrated Voltage Measurement... 14:04:25 Voltage Offset Correction: v_off_ini = 1460.407 uV, v_off_fin = 1455.170 uV 14:04:26 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 264.818 uVs 14:04:26 Integrated Voltage Measurement... 14:04:34 Voltage Offset Correction: v_off_ini = 1464.483 uV, v_off_fin = 1462.591 uV 14:04:35 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -246.270 uVs 14:04:35 Integrated Voltage Measurement... 14:04:43 Voltage Offset Correction: v_off_ini = 1477.160 uV, v_off_fin = 1482.494 uV 14:04:43 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 243.056 uVs 14:04:48 I1Y Average and RMS of Measurements VT = -252.500 +- 5.548 uVs I1Y = -50.496 +- 1.110 uTm 14:04:53 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 14:04:55 Integrated Voltage Measurement... 14:05:04 Voltage Offset Correction: v_off_ini = 1434.268 uV, v_off_fin = 1456.835 uV 14:05:04 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -27.625 uVs 14:05:04 Integrated Voltage Measurement... 14:05:13 Voltage Offset Correction: v_off_ini = 1463.156 uV, v_off_fin = 1458.944 uV 14:05:13 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 19.736 uVs 14:05:13 Integrated Voltage Measurement... 14:05:22 Voltage Offset Correction: v_off_ini = 1452.067 uV, v_off_fin = 1442.462 uV 14:05:22 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -18.774 uVs 14:05:22 Integrated Voltage Measurement... 14:05:31 Voltage Offset Correction: v_off_ini = 1433.717 uV, v_off_fin = 1451.213 uV 14:05:31 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 3.524 uVs 14:05:31 Integrated Voltage Measurement... 14:05:40 Voltage Offset Correction: v_off_ini = 1478.559 uV, v_off_fin = 1457.318 uV 14:05:40 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -12.860 uVs 14:05:40 Integrated Voltage Measurement... 14:05:49 Voltage Offset Correction: v_off_ini = 1449.754 uV, v_off_fin = 1474.053 uV 14:05:49 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -15.959 uVs 14:05:52 I2X Average and RMS of Measurements VT = -11.093 +- 10.300 uVs I2X = 1.021 +- 0.948 uTm^2 14:05:57 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 14:05:59 Integrated Voltage Measurement... 14:06:08 Voltage Offset Correction: v_off_ini = 1445.492 uV, v_off_fin = 1473.380 uV 14:06:08 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -282.898 uVs 14:06:08 Integrated Voltage Measurement... 14:06:17 Voltage Offset Correction: v_off_ini = 1450.824 uV, v_off_fin = 1484.575 uV 14:06:17 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 277.283 uVs 14:06:17 Integrated Voltage Measurement... 14:06:26 Voltage Offset Correction: v_off_ini = 1479.672 uV, v_off_fin = 1480.932 uV 14:06:26 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -247.157 uVs 14:06:26 Integrated Voltage Measurement... 14:06:35 Voltage Offset Correction: v_off_ini = 1465.652 uV, v_off_fin = 1503.755 uV 14:06:35 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 250.356 uVs 14:06:35 Integrated Voltage Measurement... 14:06:44 Voltage Offset Correction: v_off_ini = 1500.976 uV, v_off_fin = 1509.625 uV 14:06:44 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -281.028 uVs 14:06:45 Integrated Voltage Measurement... 14:06:53 Voltage Offset Correction: v_off_ini = 1504.168 uV, v_off_fin = 1482.681 uV 14:06:54 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 244.615 uVs 14:06:56 I2Y Average and RMS of Measurements VT = -263.889 +- 12.814 uVs I2Y = -24.278 +- 1.179 uTm^2 14:07:07 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 14:07:11 Integrated Voltage Measurement... 14:07:20 Voltage Offset Correction: v_off_ini = 1481.741 uV, v_off_fin = 1482.477 uV 14:07:20 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -29.842 uVs 14:07:20 Integrated Voltage Measurement... 14:07:29 Voltage Offset Correction: v_off_ini = 1482.022 uV, v_off_fin = 1496.949 uV 14:07:29 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 47.868 uVs 14:07:29 Integrated Voltage Measurement... 14:07:38 Voltage Offset Correction: v_off_ini = 1471.779 uV, v_off_fin = 1507.892 uV 14:07:38 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -66.043 uVs 14:07:38 Integrated Voltage Measurement... 14:07:47 Voltage Offset Correction: v_off_ini = 1518.361 uV, v_off_fin = 1515.827 uV 14:07:47 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 33.131 uVs 14:07:47 Integrated Voltage Measurement... 14:07:56 Voltage Offset Correction: v_off_ini = 1516.762 uV, v_off_fin = 1504.849 uV 14:07:56 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -62.180 uVs 14:07:56 Integrated Voltage Measurement... 14:08:05 Voltage Offset Correction: v_off_ini = 1491.387 uV, v_off_fin = 1480.230 uV 14:08:05 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 55.834 uVs 14:08:09 I1X Average and RMS of Measurements VT = -49.150 +- 8.233 uVs I1X = 9.830 +- 1.646 uTm 14:08:11 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 14:08:16 Integrated Voltage Measurement... 14:08:25 Voltage Offset Correction: v_off_ini = 1494.054 uV, v_off_fin = 1482.632 uV 14:08:25 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -215.321 uVs 14:08:25 Integrated Voltage Measurement... 14:08:34 Voltage Offset Correction: v_off_ini = 1487.627 uV, v_off_fin = 1511.864 uV 14:08:34 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 253.349 uVs 14:08:34 Integrated Voltage Measurement... 14:08:43 Voltage Offset Correction: v_off_ini = 1508.316 uV, v_off_fin = 1477.632 uV 14:08:43 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -227.332 uVs 14:08:43 Integrated Voltage Measurement... 14:08:52 Voltage Offset Correction: v_off_ini = 1484.071 uV, v_off_fin = 1492.881 uV 14:08:52 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 240.900 uVs 14:08:52 Integrated Voltage Measurement... 14:09:01 Voltage Offset Correction: v_off_ini = 1476.759 uV, v_off_fin = 1465.398 uV 14:09:01 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -252.972 uVs 14:09:01 Integrated Voltage Measurement... 14:09:10 Voltage Offset Correction: v_off_ini = 1477.704 uV, v_off_fin = 1487.448 uV 14:09:10 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 234.661 uVs 14:09:14 I1Y Average and RMS of Measurements VT = -237.423 +- 4.522 uVs I1Y = -47.481 +- 0.907 uTm 14:09:19 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 14:09:21 Integrated Voltage Measurement... 14:09:30 Voltage Offset Correction: v_off_ini = 1461.003 uV, v_off_fin = 1472.982 uV 14:09:30 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -10.892 uVs 14:09:30 Integrated Voltage Measurement... 14:09:39 Voltage Offset Correction: v_off_ini = 1490.798 uV, v_off_fin = 1494.877 uV 14:09:39 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 0.590 uVs 14:09:39 Integrated Voltage Measurement... 14:09:48 Voltage Offset Correction: v_off_ini = 1488.561 uV, v_off_fin = 1478.402 uV 14:09:48 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -17.623 uVs 14:09:48 Integrated Voltage Measurement... 14:09:57 Voltage Offset Correction: v_off_ini = 1475.171 uV, v_off_fin = 1502.550 uV 14:09:57 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 0.138 uVs 14:09:57 Integrated Voltage Measurement... 14:10:06 Voltage Offset Correction: v_off_ini = 1499.983 uV, v_off_fin = 1466.130 uV 14:10:06 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -21.867 uVs 14:10:06 Integrated Voltage Measurement... 14:10:15 Voltage Offset Correction: v_off_ini = 1496.671 uV, v_off_fin = 1491.516 uV 14:10:15 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -25.230 uVs 14:10:18 I2X Average and RMS of Measurements VT = -4.313 +- 4.428 uVs I2X = 0.397 +- 0.407 uTm^2 14:10:23 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 14:10:25 Integrated Voltage Measurement... 14:10:34 Voltage Offset Correction: v_off_ini = 1483.645 uV, v_off_fin = 1485.072 uV 14:10:34 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -239.046 uVs 14:10:34 Integrated Voltage Measurement... 14:10:43 Voltage Offset Correction: v_off_ini = 1467.851 uV, v_off_fin = 1508.621 uV 14:10:43 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 247.205 uVs 14:10:43 Integrated Voltage Measurement... 14:10:52 Voltage Offset Correction: v_off_ini = 1497.876 uV, v_off_fin = 1488.268 uV 14:10:52 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -240.543 uVs 14:10:53 Integrated Voltage Measurement... 14:11:01 Voltage Offset Correction: v_off_ini = 1500.847 uV, v_off_fin = 1476.698 uV 14:11:01 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 229.309 uVs 14:11:02 Integrated Voltage Measurement... 14:11:10 Voltage Offset Correction: v_off_ini = 1484.323 uV, v_off_fin = 1488.779 uV 14:11:11 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -262.438 uVs 14:11:11 Integrated Voltage Measurement... 14:11:20 Voltage Offset Correction: v_off_ini = 1483.827 uV, v_off_fin = 1484.535 uV 14:11:20 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 262.467 uVs 14:11:22 I2Y Average and RMS of Measurements VT = -246.835 +- 11.540 uVs I2Y = -22.709 +- 1.062 uTm^2 14:11:33 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 14:11:37 Integrated Voltage Measurement... 14:11:46 Voltage Offset Correction: v_off_ini = 1491.358 uV, v_off_fin = 1485.522 uV 14:11:46 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -59.556 uVs 14:11:46 Integrated Voltage Measurement... 14:11:55 Voltage Offset Correction: v_off_ini = 1474.111 uV, v_off_fin = 1479.120 uV 14:11:55 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 55.839 uVs 14:11:55 Integrated Voltage Measurement... 14:12:04 Voltage Offset Correction: v_off_ini = 1461.628 uV, v_off_fin = 1471.343 uV 14:12:04 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -28.144 uVs 14:12:04 Integrated Voltage Measurement... 14:12:13 Voltage Offset Correction: v_off_ini = 1473.324 uV, v_off_fin = 1485.748 uV 14:12:13 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 49.546 uVs 14:12:13 Integrated Voltage Measurement... 14:12:22 Voltage Offset Correction: v_off_ini = 1469.638 uV, v_off_fin = 1475.025 uV 14:12:22 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -48.829 uVs 14:12:22 Integrated Voltage Measurement... 14:12:31 Voltage Offset Correction: v_off_ini = 1473.663 uV, v_off_fin = 1490.189 uV 14:12:31 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 44.534 uVs 14:12:35 I1X Average and RMS of Measurements VT = -47.741 +- 7.733 uVs I1X = 9.548 +- 1.547 uTm 14:12:37 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 14:12:42 Integrated Voltage Measurement... 14:12:51 Voltage Offset Correction: v_off_ini = 1471.852 uV, v_off_fin = 1439.419 uV 14:12:51 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -210.954 uVs 14:12:51 Integrated Voltage Measurement... 14:13:00 Voltage Offset Correction: v_off_ini = 1469.109 uV, v_off_fin = 1492.015 uV 14:13:00 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 222.336 uVs 14:13:00 Integrated Voltage Measurement... 14:13:09 Voltage Offset Correction: v_off_ini = 1497.385 uV, v_off_fin = 1461.445 uV 14:13:09 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -218.275 uVs 14:13:09 Integrated Voltage Measurement... 14:13:18 Voltage Offset Correction: v_off_ini = 1455.583 uV, v_off_fin = 1486.978 uV 14:13:18 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 238.383 uVs 14:13:18 Integrated Voltage Measurement... 14:13:27 Voltage Offset Correction: v_off_ini = 1478.850 uV, v_off_fin = 1473.182 uV 14:13:27 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -213.108 uVs 14:13:27 Integrated Voltage Measurement... 14:13:36 Voltage Offset Correction: v_off_ini = 1492.449 uV, v_off_fin = 1485.770 uV 14:13:36 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 230.080 uVs 14:13:40 I1Y Average and RMS of Measurements VT = -222.189 +- 4.788 uVs I1Y = -44.436 +- 0.957 uTm 14:13:45 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 14:13:47 Integrated Voltage Measurement... 14:13:56 Voltage Offset Correction: v_off_ini = 1470.312 uV, v_off_fin = 1465.821 uV 14:13:56 Integrated Voltage Measurement dXYus = -1.0001 mm, dXYds = 0.0000 mm dXY = -1.0001 mm VT = 21.462 uVs 14:13:56 Integrated Voltage Measurement... 14:14:05 Voltage Offset Correction: v_off_ini = 1472.759 uV, v_off_fin = 1500.614 uV 14:14:05 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 9.470 uVs 14:14:05 Integrated Voltage Measurement... 14:14:14 Voltage Offset Correction: v_off_ini = 1488.037 uV, v_off_fin = 1480.292 uV 14:14:14 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -23.915 uVs 14:14:14 Integrated Voltage Measurement... 14:14:23 Voltage Offset Correction: v_off_ini = 1494.631 uV, v_off_fin = 1499.269 uV 14:14:23 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 7.743 uVs 14:14:24 Integrated Voltage Measurement... 14:14:32 Voltage Offset Correction: v_off_ini = 1507.035 uV, v_off_fin = 1497.428 uV 14:14:33 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 1.676 uVs 14:14:33 Integrated Voltage Measurement... 14:14:42 Voltage Offset Correction: v_off_ini = 1471.659 uV, v_off_fin = 1502.920 uV 14:14:42 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -8.909 uVs 14:14:44 I2X Average and RMS of Measurements VT = -1.514 +- 10.126 uVs I2X = 0.139 +- 0.932 uTm^2 14:14:49 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 14:14:52 Integrated Voltage Measurement... 14:15:01 Voltage Offset Correction: v_off_ini = 1520.011 uV, v_off_fin = 1505.136 uV 14:15:01 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -248.253 uVs 14:15:01 Integrated Voltage Measurement... 14:15:10 Voltage Offset Correction: v_off_ini = 1534.246 uV, v_off_fin = 1516.953 uV 14:15:10 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 238.904 uVs 14:15:10 Integrated Voltage Measurement... 14:15:19 Voltage Offset Correction: v_off_ini = 1522.804 uV, v_off_fin = 1516.280 uV 14:15:19 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -266.793 uVs 14:15:19 Integrated Voltage Measurement... 14:15:28 Voltage Offset Correction: v_off_ini = 1506.166 uV, v_off_fin = 1511.589 uV 14:15:28 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 258.720 uVs 14:15:28 Integrated Voltage Measurement... 14:15:37 Voltage Offset Correction: v_off_ini = 1530.275 uV, v_off_fin = 1514.826 uV 14:15:37 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -250.864 uVs 14:15:37 Integrated Voltage Measurement... 14:15:46 Voltage Offset Correction: v_off_ini = 1503.906 uV, v_off_fin = 1506.081 uV 14:15:46 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 246.480 uVs 14:15:48 I2Y Average and RMS of Measurements VT = -251.669 +- 8.111 uVs I2Y = -23.154 +- 0.746 uTm^2 14:15:59 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 14:16:03 Integrated Voltage Measurement... 14:16:12 Voltage Offset Correction: v_off_ini = 1482.059 uV, v_off_fin = 1494.030 uV 14:16:12 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -30.947 uVs 14:16:12 Integrated Voltage Measurement... 14:16:21 Voltage Offset Correction: v_off_ini = 1488.212 uV, v_off_fin = 1499.406 uV 14:16:21 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 69.696 uVs 14:16:21 Integrated Voltage Measurement... 14:16:30 Voltage Offset Correction: v_off_ini = 1508.234 uV, v_off_fin = 1514.898 uV 14:16:30 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -52.762 uVs 14:16:30 Integrated Voltage Measurement... 14:16:39 Voltage Offset Correction: v_off_ini = 1518.860 uV, v_off_fin = 1525.685 uV 14:16:39 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 31.577 uVs 14:16:39 Integrated Voltage Measurement... 14:16:48 Voltage Offset Correction: v_off_ini = 1520.105 uV, v_off_fin = 1503.567 uV 14:16:48 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -44.762 uVs 14:16:48 Integrated Voltage Measurement... 14:16:57 Voltage Offset Correction: v_off_ini = 1510.306 uV, v_off_fin = 1483.820 uV 14:16:57 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 16.978 uVs 14:17:01 I1X Average and RMS of Measurements VT = -41.120 +- 7.976 uVs I1X = 8.224 +- 1.595 uTm 14:17:04 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 14:17:08 Integrated Voltage Measurement... 14:17:17 Voltage Offset Correction: v_off_ini = 1476.711 uV, v_off_fin = 1475.082 uV 14:17:17 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -237.833 uVs 14:17:17 Integrated Voltage Measurement... 14:17:26 Voltage Offset Correction: v_off_ini = 1483.292 uV, v_off_fin = 1495.055 uV 14:17:26 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 230.008 uVs 14:17:26 Integrated Voltage Measurement... 14:17:35 Voltage Offset Correction: v_off_ini = 1477.174 uV, v_off_fin = 1475.131 uV 14:17:35 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -221.458 uVs 14:17:35 Integrated Voltage Measurement... 14:17:44 Voltage Offset Correction: v_off_ini = 1486.127 uV, v_off_fin = 1488.084 uV 14:17:44 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 226.634 uVs 14:17:44 Integrated Voltage Measurement... 14:17:53 Voltage Offset Correction: v_off_ini = 1485.303 uV, v_off_fin = 1447.519 uV 14:17:53 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -224.995 uVs 14:17:53 Integrated Voltage Measurement... 14:18:02 Voltage Offset Correction: v_off_ini = 1445.840 uV, v_off_fin = 1466.796 uV 14:18:02 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 243.267 uVs 14:18:06 I1Y Average and RMS of Measurements VT = -230.699 +- 4.706 uVs I1Y = -46.140 +- 0.941 uTm 14:18:11 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 14:18:14 Integrated Voltage Measurement... 14:18:23 Voltage Offset Correction: v_off_ini = 1475.187 uV, v_off_fin = 1463.568 uV 14:18:23 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -6.301 uVs 14:18:23 Integrated Voltage Measurement... 14:18:32 Voltage Offset Correction: v_off_ini = 1465.452 uV, v_off_fin = 1468.336 uV 14:18:32 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -4.482 uVs 14:18:32 Integrated Voltage Measurement... 14:18:41 Voltage Offset Correction: v_off_ini = 1460.309 uV, v_off_fin = 1444.743 uV 14:18:41 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -34.010 uVs 14:18:41 Integrated Voltage Measurement... 14:18:50 Voltage Offset Correction: v_off_ini = 1477.148 uV, v_off_fin = 1481.883 uV 14:18:50 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 20.577 uVs 14:18:50 Integrated Voltage Measurement... 14:18:59 Voltage Offset Correction: v_off_ini = 1471.930 uV, v_off_fin = 1483.180 uV 14:18:59 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -2.513 uVs 14:18:59 Integrated Voltage Measurement... 14:19:08 Voltage Offset Correction: v_off_ini = 1485.085 uV, v_off_fin = 1483.810 uV 14:19:08 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 33.357 uVs 14:19:10 I2X Average and RMS of Measurements VT = -15.379 +- 10.922 uVs I2X = 1.415 +- 1.005 uTm^2 14:19:16 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 14:19:18 Integrated Voltage Measurement... 14:19:27 Voltage Offset Correction: v_off_ini = 1479.134 uV, v_off_fin = 1485.586 uV 14:19:27 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -235.831 uVs 14:19:27 Integrated Voltage Measurement... 14:19:36 Voltage Offset Correction: v_off_ini = 1485.887 uV, v_off_fin = 1490.908 uV 14:19:36 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 235.503 uVs 14:19:36 Integrated Voltage Measurement... 14:19:45 Voltage Offset Correction: v_off_ini = 1476.616 uV, v_off_fin = 1484.290 uV 14:19:45 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -217.984 uVs 14:19:45 Integrated Voltage Measurement... 14:19:54 Voltage Offset Correction: v_off_ini = 1495.147 uV, v_off_fin = 1476.709 uV 14:19:54 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 240.821 uVs 14:19:54 Integrated Voltage Measurement... 14:20:03 Voltage Offset Correction: v_off_ini = 1473.340 uV, v_off_fin = 1471.444 uV 14:20:03 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -198.351 uVs 14:20:03 Integrated Voltage Measurement... 14:20:12 Voltage Offset Correction: v_off_ini = 1471.973 uV, v_off_fin = 1466.068 uV 14:20:12 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 232.169 uVs 14:20:15 I2Y Average and RMS of Measurements VT = -226.776 +- 8.535 uVs I2Y = -20.863 +- 0.785 uTm^2 14:20:27 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 14:20:31 Integrated Voltage Measurement... 14:20:40 Voltage Offset Correction: v_off_ini = 1465.397 uV, v_off_fin = 1478.413 uV 14:20:40 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -30.611 uVs 14:20:40 Integrated Voltage Measurement... 14:20:49 Voltage Offset Correction: v_off_ini = 1472.730 uV, v_off_fin = 1474.743 uV 14:20:49 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 28.791 uVs 14:20:49 Integrated Voltage Measurement... 14:20:58 Voltage Offset Correction: v_off_ini = 1459.677 uV, v_off_fin = 1453.215 uV 14:20:58 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -26.879 uVs 14:20:58 Integrated Voltage Measurement... 14:21:07 Voltage Offset Correction: v_off_ini = 1457.505 uV, v_off_fin = 1473.631 uV 14:21:07 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 42.858 uVs 14:21:07 Integrated Voltage Measurement... 14:21:16 Voltage Offset Correction: v_off_ini = 1452.591 uV, v_off_fin = 1474.844 uV 14:21:16 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -31.108 uVs 14:21:16 Integrated Voltage Measurement... 14:21:25 Voltage Offset Correction: v_off_ini = 1472.529 uV, v_off_fin = 1472.073 uV 14:21:25 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 37.575 uVs 14:21:29 I1X Average and RMS of Measurements VT = -32.970 +- 2.322 uVs I1X = 6.594 +- 0.465 uTm 14:21:32 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 14:21:36 Integrated Voltage Measurement... 14:21:45 Voltage Offset Correction: v_off_ini = 1486.896 uV, v_off_fin = 1468.901 uV 14:21:45 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = -240.958 uVs 14:21:45 Integrated Voltage Measurement... 14:21:54 Voltage Offset Correction: v_off_ini = 1465.856 uV, v_off_fin = 1473.061 uV 14:21:54 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 217.627 uVs 14:21:54 Integrated Voltage Measurement... 14:22:03 Voltage Offset Correction: v_off_ini = 1470.963 uV, v_off_fin = 1461.876 uV 14:22:03 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -207.119 uVs 14:22:03 Integrated Voltage Measurement... 14:22:12 Voltage Offset Correction: v_off_ini = 1469.694 uV, v_off_fin = 1481.861 uV 14:22:12 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 226.779 uVs 14:22:12 Integrated Voltage Measurement... 14:22:21 Voltage Offset Correction: v_off_ini = 1475.122 uV, v_off_fin = 1479.421 uV 14:22:21 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -212.858 uVs 14:22:21 Integrated Voltage Measurement... 14:22:30 Voltage Offset Correction: v_off_ini = 1489.169 uV, v_off_fin = 1483.755 uV 14:22:30 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 225.794 uVs 14:22:34 I1Y Average and RMS of Measurements VT = -221.856 +- 5.347 uVs I1Y = -44.371 +- 1.070 uTm 14:22:39 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 14:22:42 Integrated Voltage Measurement... 14:22:51 Voltage Offset Correction: v_off_ini = 1465.811 uV, v_off_fin = 1456.233 uV 14:22:51 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 0.262 uVs 14:22:51 Integrated Voltage Measurement... 14:23:00 Voltage Offset Correction: v_off_ini = 1470.952 uV, v_off_fin = 1488.561 uV 14:23:00 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -10.628 uVs 14:23:00 Integrated Voltage Measurement... 14:23:09 Voltage Offset Correction: v_off_ini = 1482.012 uV, v_off_fin = 1476.830 uV 14:23:09 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -28.961 uVs 14:23:09 Integrated Voltage Measurement... 14:23:18 Voltage Offset Correction: v_off_ini = 1505.310 uV, v_off_fin = 1480.710 uV 14:23:18 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 27.809 uVs 14:23:18 Integrated Voltage Measurement... 14:23:27 Voltage Offset Correction: v_off_ini = 1474.155 uV, v_off_fin = 1472.305 uV 14:23:27 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 12.799 uVs 14:23:27 Integrated Voltage Measurement... 14:23:36 Voltage Offset Correction: v_off_ini = 1476.409 uV, v_off_fin = 1472.334 uV 14:23:36 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 0.309 uVs 14:23:38 I2X Average and RMS of Measurements VT = -5.565 +- 16.139 uVs I2X = 0.512 +- 1.485 uTm^2 14:23:43 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 14:23:46 Integrated Voltage Measurement... 14:23:55 Voltage Offset Correction: v_off_ini = 1470.180 uV, v_off_fin = 1488.066 uV 14:23:55 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0001 mm dXY = 1.0000 mm VT = -206.074 uVs 14:23:55 Integrated Voltage Measurement... 14:24:04 Voltage Offset Correction: v_off_ini = 1491.095 uV, v_off_fin = 1504.602 uV 14:24:04 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 215.841 uVs 14:24:04 Integrated Voltage Measurement... 14:24:13 Voltage Offset Correction: v_off_ini = 1487.217 uV, v_off_fin = 1494.583 uV 14:24:13 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -214.888 uVs 14:24:13 Integrated Voltage Measurement... 14:24:22 Voltage Offset Correction: v_off_ini = 1500.994 uV, v_off_fin = 1497.035 uV 14:24:22 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 231.357 uVs 14:24:22 Integrated Voltage Measurement... 14:24:31 Voltage Offset Correction: v_off_ini = 1516.989 uV, v_off_fin = 1509.032 uV 14:24:31 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -217.560 uVs 14:24:31 Integrated Voltage Measurement... 14:24:40 Voltage Offset Correction: v_off_ini = 1529.845 uV, v_off_fin = 1541.276 uV 14:24:40 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 213.777 uVs 14:24:42 I2Y Average and RMS of Measurements VT = -216.583 +- 5.008 uVs I2Y = -19.926 +- 0.461 uTm^2 14:24:54 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 14:24:59 Integrated Voltage Measurement... 14:25:07 Voltage Offset Correction: v_off_ini = 1540.111 uV, v_off_fin = 1528.531 uV 14:25:07 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -61.498 uVs 14:25:08 Integrated Voltage Measurement... 14:25:16 Voltage Offset Correction: v_off_ini = 1539.529 uV, v_off_fin = 1544.658 uV 14:25:16 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = 25.214 uVs 14:25:17 Integrated Voltage Measurement... 14:25:25 Voltage Offset Correction: v_off_ini = 1537.741 uV, v_off_fin = 1545.762 uV 14:25:25 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -17.958 uVs 14:25:26 Integrated Voltage Measurement... 14:25:34 Voltage Offset Correction: v_off_ini = 1543.674 uV, v_off_fin = 1586.184 uV 14:25:34 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 41.102 uVs 14:25:35 Integrated Voltage Measurement... 14:25:43 Voltage Offset Correction: v_off_ini = 1552.965 uV, v_off_fin = 1539.129 uV 14:25:44 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -32.908 uVs 14:25:44 Integrated Voltage Measurement... 14:25:52 Voltage Offset Correction: v_off_ini = 1524.988 uV, v_off_fin = 1552.888 uV 14:25:53 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 36.053 uVs 14:25:57 I1X Average and RMS of Measurements VT = -35.789 +- 5.720 uVs I1X = 7.158 +- 1.144 uTm 14:25:59 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 14:26:03 Integrated Voltage Measurement... 14:26:12 Voltage Offset Correction: v_off_ini = 1541.682 uV, v_off_fin = 1515.850 uV 14:26:12 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -211.463 uVs 14:26:12 Integrated Voltage Measurement... 14:26:21 Voltage Offset Correction: v_off_ini = 1535.742 uV, v_off_fin = 1529.866 uV 14:26:21 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 215.962 uVs 14:26:21 Integrated Voltage Measurement... 14:26:30 Voltage Offset Correction: v_off_ini = 1520.363 uV, v_off_fin = 1511.361 uV 14:26:30 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -195.980 uVs 14:26:30 Integrated Voltage Measurement... 14:26:39 Voltage Offset Correction: v_off_ini = 1513.361 uV, v_off_fin = 1539.957 uV 14:26:39 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 212.982 uVs 14:26:39 Integrated Voltage Measurement... 14:26:48 Voltage Offset Correction: v_off_ini = 1551.119 uV, v_off_fin = 1533.459 uV 14:26:48 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -207.831 uVs 14:26:48 Integrated Voltage Measurement... 14:26:57 Voltage Offset Correction: v_off_ini = 1535.401 uV, v_off_fin = 1554.767 uV 14:26:57 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 206.889 uVs 14:27:02 I1Y Average and RMS of Measurements VT = -208.518 +- 3.856 uVs I1Y = -41.700 +- 0.771 uTm 14:27:07 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 14:27:09 Integrated Voltage Measurement... 14:27:18 Voltage Offset Correction: v_off_ini = 1543.689 uV, v_off_fin = 1537.614 uV 14:27:18 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -27.990 uVs 14:27:18 Integrated Voltage Measurement... 14:27:27 Voltage Offset Correction: v_off_ini = 1533.687 uV, v_off_fin = 1535.681 uV 14:27:27 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 14.179 uVs 14:27:27 Integrated Voltage Measurement... 14:27:36 Voltage Offset Correction: v_off_ini = 1517.722 uV, v_off_fin = 1506.795 uV 14:27:36 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -5.345 uVs 14:27:36 Integrated Voltage Measurement... 14:27:45 Voltage Offset Correction: v_off_ini = 1530.001 uV, v_off_fin = 1537.964 uV 14:27:45 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 19.287 uVs 14:27:45 Integrated Voltage Measurement... 14:27:54 Voltage Offset Correction: v_off_ini = 1539.986 uV, v_off_fin = 1528.669 uV 14:27:54 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -2.298 uVs 14:27:54 Integrated Voltage Measurement... 14:28:03 Voltage Offset Correction: v_off_ini = 1523.487 uV, v_off_fin = 1537.664 uV 14:28:03 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 18.551 uVs 14:28:06 I2X Average and RMS of Measurements VT = -14.608 +- 4.644 uVs I2X = 1.344 +- 0.427 uTm^2 14:28:11 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 14:28:13 Integrated Voltage Measurement... 14:28:22 Voltage Offset Correction: v_off_ini = 1531.894 uV, v_off_fin = 1557.772 uV 14:28:22 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = -211.474 uVs 14:28:22 Integrated Voltage Measurement... 14:28:31 Voltage Offset Correction: v_off_ini = 1521.843 uV, v_off_fin = 1528.140 uV 14:28:31 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 208.295 uVs 14:28:32 Integrated Voltage Measurement... 14:28:40 Voltage Offset Correction: v_off_ini = 1537.779 uV, v_off_fin = 1555.403 uV 14:28:41 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -182.905 uVs 14:28:41 Integrated Voltage Measurement... 14:28:50 Voltage Offset Correction: v_off_ini = 1551.718 uV, v_off_fin = 1533.260 uV 14:28:50 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 219.802 uVs 14:28:50 Integrated Voltage Measurement... 14:28:59 Voltage Offset Correction: v_off_ini = 1543.211 uV, v_off_fin = 1537.367 uV 14:28:59 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -202.233 uVs 14:28:59 Integrated Voltage Measurement... 14:29:08 Voltage Offset Correction: v_off_ini = 1550.095 uV, v_off_fin = 1544.390 uV 14:29:08 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 213.314 uVs 14:29:10 I2Y Average and RMS of Measurements VT = -206.337 +- 3.628 uVs I2Y = -18.983 +- 0.334 uTm^2 14:29:25 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 14:29:30 Integrated Voltage Measurement... 14:29:38 Voltage Offset Correction: v_off_ini = 1547.947 uV, v_off_fin = 1543.056 uV 14:29:39 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -40.959 uVs 14:29:39 Integrated Voltage Measurement... 14:29:48 Voltage Offset Correction: v_off_ini = 1519.616 uV, v_off_fin = 1528.110 uV 14:29:48 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 56.328 uVs 14:29:48 Integrated Voltage Measurement... 14:29:57 Voltage Offset Correction: v_off_ini = 1537.139 uV, v_off_fin = 1536.662 uV 14:29:57 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -29.713 uVs 14:29:57 Integrated Voltage Measurement... 14:30:06 Voltage Offset Correction: v_off_ini = 1551.914 uV, v_off_fin = 1557.538 uV 14:30:06 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 40.334 uVs 14:30:06 Integrated Voltage Measurement... 14:30:15 Voltage Offset Correction: v_off_ini = 1546.406 uV, v_off_fin = 1542.869 uV 14:30:15 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -45.247 uVs 14:30:15 Integrated Voltage Measurement... 14:30:24 Voltage Offset Correction: v_off_ini = 1542.240 uV, v_off_fin = 1547.551 uV 14:30:24 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 22.663 uVs 14:30:28 I1X Average and RMS of Measurements VT = -39.207 +- 6.687 uVs I1X = 7.842 +- 1.337 uTm 14:30:30 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 14:30:35 Integrated Voltage Measurement... 14:30:43 Voltage Offset Correction: v_off_ini = 1551.439 uV, v_off_fin = 1502.944 uV 14:30:43 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -174.389 uVs 14:30:44 Integrated Voltage Measurement... 14:30:52 Voltage Offset Correction: v_off_ini = 1555.977 uV, v_off_fin = 1586.526 uV 14:30:52 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 208.386 uVs 14:30:52 Integrated Voltage Measurement... 14:31:01 Voltage Offset Correction: v_off_ini = 1546.424 uV, v_off_fin = 1533.516 uV 14:31:01 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -194.573 uVs 14:31:01 Integrated Voltage Measurement... 14:31:10 Voltage Offset Correction: v_off_ini = 1516.412 uV, v_off_fin = 1538.334 uV 14:31:10 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 199.661 uVs 14:31:10 Integrated Voltage Measurement... 14:31:19 Voltage Offset Correction: v_off_ini = 1527.455 uV, v_off_fin = 1527.221 uV 14:31:19 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -191.370 uVs 14:31:19 Integrated Voltage Measurement... 14:31:28 Voltage Offset Correction: v_off_ini = 1521.927 uV, v_off_fin = 1552.835 uV 14:31:28 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 213.487 uVs 14:31:33 I1Y Average and RMS of Measurements VT = -196.977 +- 4.509 uVs I1Y = -39.393 +- 0.902 uTm 14:31:38 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 14:31:40 Integrated Voltage Measurement... 14:31:49 Voltage Offset Correction: v_off_ini = 1518.277 uV, v_off_fin = 1528.569 uV 14:31:49 Integrated Voltage Measurement dXYus = -1.0001 mm, dXYds = 0.0000 mm dXY = -1.0001 mm VT = -7.381 uVs 14:31:49 Integrated Voltage Measurement... 14:31:58 Voltage Offset Correction: v_off_ini = 1532.712 uV, v_off_fin = 1545.526 uV 14:31:58 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -18.974 uVs 14:31:58 Integrated Voltage Measurement... 14:32:07 Voltage Offset Correction: v_off_ini = 1520.146 uV, v_off_fin = 1516.375 uV 14:32:07 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -38.840 uVs 14:32:07 Integrated Voltage Measurement... 14:32:16 Voltage Offset Correction: v_off_ini = 1536.385 uV, v_off_fin = 1540.118 uV 14:32:16 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 11.199 uVs 14:32:16 Integrated Voltage Measurement... 14:32:25 Voltage Offset Correction: v_off_ini = 1528.031 uV, v_off_fin = 1530.145 uV 14:32:25 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -10.698 uVs 14:32:25 Integrated Voltage Measurement... 14:32:34 Voltage Offset Correction: v_off_ini = 1544.285 uV, v_off_fin = 1544.285 uV 14:32:34 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 18.526 uVs 14:32:37 I2X Average and RMS of Measurements VT = -11.278 +- 12.800 uVs I2X = 1.038 +- 1.178 uTm^2 14:32:42 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 14:32:44 Integrated Voltage Measurement... 14:32:53 Voltage Offset Correction: v_off_ini = 1511.335 uV, v_off_fin = 1523.120 uV 14:32:54 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = -185.237 uVs 14:32:54 Integrated Voltage Measurement... 14:33:03 Voltage Offset Correction: v_off_ini = 1523.371 uV, v_off_fin = 1533.157 uV 14:33:03 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 200.768 uVs 14:33:03 Integrated Voltage Measurement... 14:33:12 Voltage Offset Correction: v_off_ini = 1525.376 uV, v_off_fin = 1535.932 uV 14:33:12 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -181.903 uVs 14:33:12 Integrated Voltage Measurement... 14:33:21 Voltage Offset Correction: v_off_ini = 1542.234 uV, v_off_fin = 1544.982 uV 14:33:21 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 191.827 uVs 14:33:21 Integrated Voltage Measurement... 14:33:30 Voltage Offset Correction: v_off_ini = 1541.340 uV, v_off_fin = 1559.920 uV 14:33:30 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -158.682 uVs 14:33:30 Integrated Voltage Measurement... 14:33:39 Voltage Offset Correction: v_off_ini = 1540.614 uV, v_off_fin = 1557.597 uV 14:33:39 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 183.187 uVs 14:33:41 I2Y Average and RMS of Measurements VT = -183.601 +- 9.300 uVs I2Y = -16.892 +- 0.856 uTm^2 14:33:56 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 14:34:01 Integrated Voltage Measurement... 14:34:10 Voltage Offset Correction: v_off_ini = 1566.189 uV, v_off_fin = 1567.666 uV 14:34:10 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -52.957 uVs 14:34:10 Integrated Voltage Measurement... 14:34:19 Voltage Offset Correction: v_off_ini = 1547.847 uV, v_off_fin = 1537.969 uV 14:34:19 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 43.257 uVs 14:34:19 Integrated Voltage Measurement... 14:34:28 Voltage Offset Correction: v_off_ini = 1552.534 uV, v_off_fin = 1576.439 uV 14:34:28 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -35.867 uVs 14:34:28 Integrated Voltage Measurement... 14:34:37 Voltage Offset Correction: v_off_ini = 1563.949 uV, v_off_fin = 1588.856 uV 14:34:37 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 42.900 uVs 14:34:37 Integrated Voltage Measurement... 14:34:46 Voltage Offset Correction: v_off_ini = 1575.661 uV, v_off_fin = 1582.662 uV 14:34:46 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -27.783 uVs 14:34:46 Integrated Voltage Measurement... 14:34:55 Voltage Offset Correction: v_off_ini = 1605.993 uV, v_off_fin = 1592.652 uV 14:34:55 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 41.076 uVs 14:34:59 I1X Average and RMS of Measurements VT = -40.640 +- 5.654 uVs I1X = 8.128 +- 1.131 uTm 14:35:01 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 14:46:01 Integrated Voltage Measurement... 14:46:10 Voltage Offset Correction: v_off_ini = 1527.248 uV, v_off_fin = 1531.134 uV 14:46:10 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -174.080 uVs 14:46:10 Integrated Voltage Measurement... 14:46:19 Voltage Offset Correction: v_off_ini = 1541.472 uV, v_off_fin = 1547.816 uV 14:46:19 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 178.447 uVs 14:46:19 Integrated Voltage Measurement... 14:46:28 Voltage Offset Correction: v_off_ini = 1523.760 uV, v_off_fin = 1534.327 uV 14:46:28 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -190.235 uVs 14:46:28 Integrated Voltage Measurement... 14:46:37 Voltage Offset Correction: v_off_ini = 1513.181 uV, v_off_fin = 1524.186 uV 14:46:37 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 193.548 uVs 14:46:37 Integrated Voltage Measurement... 14:46:46 Voltage Offset Correction: v_off_ini = 1528.982 uV, v_off_fin = 1525.591 uV 14:46:46 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -173.368 uVs 14:46:46 Integrated Voltage Measurement... 14:46:55 Voltage Offset Correction: v_off_ini = 1529.143 uV, v_off_fin = 1551.246 uV 14:46:55 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 200.485 uVs 14:46:59 I1Y Average and RMS of Measurements VT = -185.027 +- 6.520 uVs I1Y = -37.003 +- 1.302 uTm 14:47:05 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 14:47:07 Integrated Voltage Measurement... 14:47:16 Voltage Offset Correction: v_off_ini = 1519.749 uV, v_off_fin = 1515.893 uV 14:47:16 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 6.660 uVs 14:47:16 Integrated Voltage Measurement... 14:47:25 Voltage Offset Correction: v_off_ini = 1547.045 uV, v_off_fin = 1555.382 uV 14:47:25 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 8.526 uVs 14:47:25 Integrated Voltage Measurement... 14:47:34 Voltage Offset Correction: v_off_ini = 1566.737 uV, v_off_fin = 1526.153 uV 14:47:34 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -7.055 uVs 14:47:34 Integrated Voltage Measurement... 14:47:43 Voltage Offset Correction: v_off_ini = 1530.048 uV, v_off_fin = 1505.935 uV 14:47:43 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 12.394 uVs 14:47:43 Integrated Voltage Measurement... 14:47:52 Voltage Offset Correction: v_off_ini = 1505.884 uV, v_off_fin = 1526.869 uV 14:47:52 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -3.779 uVs 14:47:52 Integrated Voltage Measurement... 14:48:01 Voltage Offset Correction: v_off_ini = 1519.990 uV, v_off_fin = 1524.379 uV 14:48:01 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 15.076 uVs 14:48:04 I2X Average and RMS of Measurements VT = -6.695 +- 4.076 uVs I2X = 0.616 +- 0.375 uTm^2 14:48:09 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 14:48:11 Integrated Voltage Measurement... 14:48:20 Voltage Offset Correction: v_off_ini = 1505.192 uV, v_off_fin = 1516.790 uV 14:48:20 Integrated Voltage Measurement dXYus = 1.0001 mm, dXYds = 0.0000 mm dXY = 1.0001 mm VT = -169.991 uVs 14:48:20 Integrated Voltage Measurement... 14:48:29 Voltage Offset Correction: v_off_ini = 1522.603 uV, v_off_fin = 1535.051 uV 14:48:29 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 189.073 uVs 14:48:29 Integrated Voltage Measurement... 14:48:38 Voltage Offset Correction: v_off_ini = 1546.278 uV, v_off_fin = 1542.998 uV 14:48:38 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -184.260 uVs 14:48:38 Integrated Voltage Measurement... 14:48:47 Voltage Offset Correction: v_off_ini = 1547.774 uV, v_off_fin = 1574.699 uV 14:48:47 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 180.924 uVs 14:48:47 Integrated Voltage Measurement... 14:48:56 Voltage Offset Correction: v_off_ini = 1558.301 uV, v_off_fin = 1562.321 uV 14:48:56 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -162.011 uVs 14:48:56 Integrated Voltage Measurement... 14:49:05 Voltage Offset Correction: v_off_ini = 1559.086 uV, v_off_fin = 1555.481 uV 14:49:05 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 193.754 uVs 14:49:08 I2Y Average and RMS of Measurements VT = -180.002 +- 1.951 uVs I2Y = -16.560 +- 0.180 uTm^2 14:49:23 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 14:49:27 Integrated Voltage Measurement... 14:49:36 Voltage Offset Correction: v_off_ini = 1515.014 uV, v_off_fin = 1524.767 uV 14:49:36 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -45.639 uVs 14:49:36 Integrated Voltage Measurement... 14:49:45 Voltage Offset Correction: v_off_ini = 1556.912 uV, v_off_fin = 1541.900 uV 14:49:45 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 48.947 uVs 14:49:45 Integrated Voltage Measurement... 14:49:54 Voltage Offset Correction: v_off_ini = 1525.804 uV, v_off_fin = 1540.213 uV 14:49:54 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -14.873 uVs 14:49:54 Integrated Voltage Measurement... 14:50:03 Voltage Offset Correction: v_off_ini = 1534.961 uV, v_off_fin = 1560.837 uV 14:50:03 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 42.333 uVs 14:50:03 Integrated Voltage Measurement... 14:50:12 Voltage Offset Correction: v_off_ini = 1542.964 uV, v_off_fin = 1532.909 uV 14:50:12 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.4999 mm dXY = -0.4999 mm VT = -52.768 uVs 14:50:12 Integrated Voltage Measurement... 14:50:21 Voltage Offset Correction: v_off_ini = 1531.851 uV, v_off_fin = 1535.991 uV 14:50:21 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = 20.163 uVs 14:50:25 I1X Average and RMS of Measurements VT = -37.454 +- 7.662 uVs I1X = 7.491 +- 1.532 uTm 14:50:28 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 14:50:32 Integrated Voltage Measurement... 14:50:41 Voltage Offset Correction: v_off_ini = 1529.981 uV, v_off_fin = 1517.665 uV 14:50:41 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -180.788 uVs 14:50:41 Integrated Voltage Measurement... 14:50:50 Voltage Offset Correction: v_off_ini = 1512.522 uV, v_off_fin = 1531.552 uV 14:50:50 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 184.105 uVs 14:50:50 Integrated Voltage Measurement... 14:50:59 Voltage Offset Correction: v_off_ini = 1549.001 uV, v_off_fin = 1535.060 uV 14:50:59 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -177.127 uVs 14:50:59 Integrated Voltage Measurement... 14:51:08 Voltage Offset Correction: v_off_ini = 1524.310 uV, v_off_fin = 1521.052 uV 14:51:08 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 191.285 uVs 14:51:08 Integrated Voltage Measurement... 14:51:17 Voltage Offset Correction: v_off_ini = 1542.111 uV, v_off_fin = 1534.841 uV 14:51:17 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -174.483 uVs 14:51:17 Integrated Voltage Measurement... 14:51:26 Voltage Offset Correction: v_off_ini = 1533.181 uV, v_off_fin = 1554.511 uV 14:51:26 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 188.611 uVs 14:51:30 I1Y Average and RMS of Measurements VT = -182.733 +- 1.104 uVs I1Y = -36.543 +- 0.221 uTm 14:51:35 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 14:51:38 Integrated Voltage Measurement... 14:51:47 Voltage Offset Correction: v_off_ini = 1561.456 uV, v_off_fin = 1557.432 uV 14:51:47 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -23.801 uVs 14:51:47 Integrated Voltage Measurement... 14:51:56 Voltage Offset Correction: v_off_ini = 1567.826 uV, v_off_fin = 1564.455 uV 14:51:56 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 30.964 uVs 14:51:56 Integrated Voltage Measurement... 14:52:05 Voltage Offset Correction: v_off_ini = 1548.304 uV, v_off_fin = 1533.651 uV 14:52:05 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -21.239 uVs 14:52:05 Integrated Voltage Measurement... 14:52:14 Voltage Offset Correction: v_off_ini = 1548.464 uV, v_off_fin = 1575.921 uV 14:52:14 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 45.861 uVs 14:52:14 Integrated Voltage Measurement... 14:52:23 Voltage Offset Correction: v_off_ini = 1580.173 uV, v_off_fin = 1561.653 uV 14:52:23 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -32.235 uVs 14:52:23 Integrated Voltage Measurement... 14:52:32 Voltage Offset Correction: v_off_ini = 1564.943 uV, v_off_fin = 1574.397 uV 14:52:32 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 13.887 uVs 14:52:34 I2X Average and RMS of Measurements VT = -27.998 +- 4.304 uVs I2X = 2.576 +- 0.396 uTm^2 14:52:39 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 14:52:42 Integrated Voltage Measurement... 14:52:51 Voltage Offset Correction: v_off_ini = 1533.188 uV, v_off_fin = 1559.749 uV 14:52:51 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -165.079 uVs 14:52:51 Integrated Voltage Measurement... 14:53:00 Voltage Offset Correction: v_off_ini = 1547.899 uV, v_off_fin = 1554.985 uV 14:53:00 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 164.837 uVs 14:53:00 Integrated Voltage Measurement... 14:53:09 Voltage Offset Correction: v_off_ini = 1569.787 uV, v_off_fin = 1576.401 uV 14:53:09 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -163.204 uVs 14:53:09 Integrated Voltage Measurement... 14:53:18 Voltage Offset Correction: v_off_ini = 1557.819 uV, v_off_fin = 1558.612 uV 14:53:18 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 168.718 uVs 14:53:18 Integrated Voltage Measurement... 14:53:27 Voltage Offset Correction: v_off_ini = 1561.964 uV, v_off_fin = 1567.493 uV 14:53:27 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -161.032 uVs 14:53:27 Integrated Voltage Measurement... 14:53:36 Voltage Offset Correction: v_off_ini = 1562.242 uV, v_off_fin = 1562.241 uV 14:53:36 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 163.897 uVs 14:53:38 I2Y Average and RMS of Measurements VT = -164.461 +- 1.470 uVs I2Y = -15.130 +- 0.135 uTm^2 14:53:53 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 14:53:57 Integrated Voltage Measurement... 14:54:06 Voltage Offset Correction: v_off_ini = 1597.697 uV, v_off_fin = 1585.053 uV 14:54:06 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -66.063 uVs 14:54:06 Integrated Voltage Measurement... 14:54:15 Voltage Offset Correction: v_off_ini = 1604.225 uV, v_off_fin = 1607.589 uV 14:54:15 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 45.465 uVs 14:54:15 Integrated Voltage Measurement... 14:54:24 Voltage Offset Correction: v_off_ini = 1594.093 uV, v_off_fin = 1605.357 uV 14:54:24 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -24.826 uVs 14:54:24 Integrated Voltage Measurement... 14:54:33 Voltage Offset Correction: v_off_ini = 1600.611 uV, v_off_fin = 1591.076 uV 14:54:33 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 56.286 uVs 14:54:33 Integrated Voltage Measurement... 14:54:42 Voltage Offset Correction: v_off_ini = 1586.630 uV, v_off_fin = 1603.153 uV 14:54:42 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -61.283 uVs 14:54:42 Integrated Voltage Measurement... 14:54:51 Voltage Offset Correction: v_off_ini = 1620.197 uV, v_off_fin = 1625.642 uV 14:54:51 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 59.736 uVs 14:54:55 I1X Average and RMS of Measurements VT = -52.276 +- 8.511 uVs I1X = 10.456 +- 1.702 uTm 14:54:58 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 14:55:02 Integrated Voltage Measurement... 14:55:11 Voltage Offset Correction: v_off_ini = 1601.283 uV, v_off_fin = 1581.659 uV 14:55:11 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = -202.470 uVs 14:55:11 Integrated Voltage Measurement... 14:55:20 Voltage Offset Correction: v_off_ini = 1639.997 uV, v_off_fin = 1607.721 uV 14:55:20 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 161.341 uVs 14:55:20 Integrated Voltage Measurement... 14:55:29 Voltage Offset Correction: v_off_ini = 1600.267 uV, v_off_fin = 1615.695 uV 14:55:29 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -166.046 uVs 14:55:29 Integrated Voltage Measurement... 14:55:38 Voltage Offset Correction: v_off_ini = 1613.213 uV, v_off_fin = 1633.985 uV 14:55:38 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 185.962 uVs 14:55:38 Integrated Voltage Measurement... 14:55:47 Voltage Offset Correction: v_off_ini = 1618.995 uV, v_off_fin = 1623.822 uV 14:55:47 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -169.270 uVs 14:55:47 Integrated Voltage Measurement... 14:55:56 Voltage Offset Correction: v_off_ini = 1632.334 uV, v_off_fin = 1626.927 uV 14:55:56 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 190.307 uVs 14:56:00 I1Y Average and RMS of Measurements VT = -179.233 +- 2.441 uVs I1Y = -35.845 +- 0.490 uTm 14:56:05 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 14:56:08 Integrated Voltage Measurement... 14:56:17 Voltage Offset Correction: v_off_ini = 1611.274 uV, v_off_fin = 1618.821 uV 14:56:17 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 0.792 uVs 14:56:17 Integrated Voltage Measurement... 14:56:26 Voltage Offset Correction: v_off_ini = 1609.907 uV, v_off_fin = 1616.065 uV 14:56:26 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 48.246 uVs 14:56:26 Integrated Voltage Measurement... 14:56:35 Voltage Offset Correction: v_off_ini = 1634.502 uV, v_off_fin = 1632.547 uV 14:56:35 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -13.320 uVs 14:56:35 Integrated Voltage Measurement... 14:56:44 Voltage Offset Correction: v_off_ini = 1662.018 uV, v_off_fin = 1661.182 uV 14:56:44 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 31.667 uVs 14:56:44 Integrated Voltage Measurement... 14:56:53 Voltage Offset Correction: v_off_ini = 1625.658 uV, v_off_fin = 1626.391 uV 14:56:53 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -43.846 uVs 14:56:53 Integrated Voltage Measurement... 14:57:02 Voltage Offset Correction: v_off_ini = 1664.375 uV, v_off_fin = 1687.733 uV 14:57:02 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 13.923 uVs 14:57:04 I2X Average and RMS of Measurements VT = -25.035 +- 2.768 uVs I2X = 2.303 +- 0.255 uTm^2 14:57:10 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 14:57:12 Integrated Voltage Measurement... 14:57:21 Voltage Offset Correction: v_off_ini = 1673.794 uV, v_off_fin = 1685.877 uV 14:57:21 Integrated Voltage Measurement dXYus = 1.0001 mm, dXYds = 0.0000 mm dXY = 1.0001 mm VT = -180.842 uVs 14:57:21 Integrated Voltage Measurement... 14:57:30 Voltage Offset Correction: v_off_ini = 1708.599 uV, v_off_fin = 1698.357 uV 14:57:30 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 150.185 uVs 14:57:30 Integrated Voltage Measurement... 14:57:39 Voltage Offset Correction: v_off_ini = 1711.557 uV, v_off_fin = 1712.567 uV 14:57:39 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -161.544 uVs 14:57:39 Integrated Voltage Measurement... 14:57:48 Voltage Offset Correction: v_off_ini = 1731.819 uV, v_off_fin = 1733.077 uV 14:57:48 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 169.112 uVs 14:57:48 Integrated Voltage Measurement... 14:57:57 Voltage Offset Correction: v_off_ini = 1718.318 uV, v_off_fin = 1718.464 uV 14:57:57 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -156.134 uVs 14:57:57 Integrated Voltage Measurement... 14:58:06 Voltage Offset Correction: v_off_ini = 1721.090 uV, v_off_fin = 1734.047 uV 14:58:06 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 180.689 uVs 14:58:09 I2Y Average and RMS of Measurements VT = -166.418 +- 1.412 uVs I2Y = -15.310 +- 0.130 uTm^2 14:58:23 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 14:58:28 Integrated Voltage Measurement... 14:58:37 Voltage Offset Correction: v_off_ini = 1735.269 uV, v_off_fin = 1744.124 uV 14:58:37 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.4999 mm dXY = -0.4999 mm VT = -65.785 uVs 14:58:37 Integrated Voltage Measurement... 14:58:46 Voltage Offset Correction: v_off_ini = 1743.516 uV, v_off_fin = 1737.974 uV 14:58:46 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.4999 mm dXY = 0.4999 mm VT = 69.764 uVs 14:58:46 Integrated Voltage Measurement... 14:58:55 Voltage Offset Correction: v_off_ini = 1729.387 uV, v_off_fin = 1722.193 uV 14:58:55 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -56.175 uVs 14:58:55 Integrated Voltage Measurement... 14:59:04 Voltage Offset Correction: v_off_ini = 1744.062 uV, v_off_fin = 1742.226 uV 14:59:04 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 60.819 uVs 14:59:04 Integrated Voltage Measurement... 14:59:13 Voltage Offset Correction: v_off_ini = 1758.966 uV, v_off_fin = 1732.767 uV 14:59:13 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -70.660 uVs 14:59:13 Integrated Voltage Measurement... 14:59:22 Voltage Offset Correction: v_off_ini = 1719.142 uV, v_off_fin = 1712.485 uV 14:59:22 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 83.300 uVs 14:59:26 I1X Average and RMS of Measurements VT = -67.750 +- 7.546 uVs I1X = 13.551 +- 1.509 uTm 14:59:28 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 14:59:33 Integrated Voltage Measurement... 14:59:42 Voltage Offset Correction: v_off_ini = 1725.007 uV, v_off_fin = 1723.573 uV 14:59:42 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -173.529 uVs 14:59:42 Integrated Voltage Measurement... 14:59:51 Voltage Offset Correction: v_off_ini = 1736.120 uV, v_off_fin = 1758.629 uV 14:59:51 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 159.447 uVs 14:59:51 Integrated Voltage Measurement... 15:00:00 Voltage Offset Correction: v_off_ini = 1738.812 uV, v_off_fin = 1721.931 uV 15:00:00 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -193.220 uVs 15:00:00 Integrated Voltage Measurement... 15:00:09 Voltage Offset Correction: v_off_ini = 1697.832 uV, v_off_fin = 1710.189 uV 15:00:09 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 175.918 uVs 15:00:09 Integrated Voltage Measurement... 15:00:18 Voltage Offset Correction: v_off_ini = 1721.670 uV, v_off_fin = 1707.182 uV 15:00:18 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5002 mm dXY = 0.5001 mm VT = -153.172 uVs 15:00:18 Integrated Voltage Measurement... 15:00:26 Voltage Offset Correction: v_off_ini = 1724.125 uV, v_off_fin = 1737.166 uV 15:00:27 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 173.690 uVs 15:00:31 I1Y Average and RMS of Measurements VT = -171.496 +- 9.328 uVs I1Y = -34.296 +- 1.865 uTm 15:00:36 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 15:00:38 Integrated Voltage Measurement... 15:00:47 Voltage Offset Correction: v_off_ini = 1736.158 uV, v_off_fin = 1720.988 uV 15:00:47 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -36.486 uVs 15:00:47 Integrated Voltage Measurement... 15:00:56 Voltage Offset Correction: v_off_ini = 1741.360 uV, v_off_fin = 1737.330 uV 15:00:56 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 39.518 uVs 15:00:56 Integrated Voltage Measurement... 15:01:05 Voltage Offset Correction: v_off_ini = 1736.010 uV, v_off_fin = 1731.552 uV 15:01:05 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -30.178 uVs 15:01:05 Integrated Voltage Measurement... 15:01:14 Voltage Offset Correction: v_off_ini = 1736.413 uV, v_off_fin = 1738.836 uV 15:01:14 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 45.019 uVs 15:01:14 Integrated Voltage Measurement... 15:01:23 Voltage Offset Correction: v_off_ini = 1708.981 uV, v_off_fin = 1694.678 uV 15:01:23 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -20.470 uVs 15:01:24 Integrated Voltage Measurement... 15:01:32 Voltage Offset Correction: v_off_ini = 1733.133 uV, v_off_fin = 1774.130 uV 15:01:32 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 31.501 uVs 15:01:35 I2X Average and RMS of Measurements VT = -33.862 +- 5.572 uVs I2X = 3.115 +- 0.513 uTm^2 15:01:40 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 15:01:43 Integrated Voltage Measurement... 15:01:52 Voltage Offset Correction: v_off_ini = 1730.034 uV, v_off_fin = 1727.194 uV 15:01:52 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -152.720 uVs 15:01:52 Integrated Voltage Measurement... 15:02:01 Voltage Offset Correction: v_off_ini = 1726.689 uV, v_off_fin = 1720.110 uV 15:02:01 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 152.306 uVs 15:02:01 Integrated Voltage Measurement... 15:02:10 Voltage Offset Correction: v_off_ini = 1733.686 uV, v_off_fin = 1735.545 uV 15:02:10 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -166.502 uVs 15:02:10 Integrated Voltage Measurement... 15:02:19 Voltage Offset Correction: v_off_ini = 1743.881 uV, v_off_fin = 1724.904 uV 15:02:19 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 166.213 uVs 15:02:19 Integrated Voltage Measurement... 15:02:28 Voltage Offset Correction: v_off_ini = 1725.109 uV, v_off_fin = 1721.534 uV 15:02:28 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -152.023 uVs 15:02:28 Integrated Voltage Measurement... 15:02:37 Voltage Offset Correction: v_off_ini = 1702.397 uV, v_off_fin = 1717.804 uV 15:02:37 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 181.003 uVs 15:02:39 I2Y Average and RMS of Measurements VT = -161.795 +- 6.563 uVs I2Y = -14.885 +- 0.604 uTm^2 15:02:54 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 15:02:58 Integrated Voltage Measurement... 15:03:07 Voltage Offset Correction: v_off_ini = 1731.370 uV, v_off_fin = 1721.195 uV 15:03:07 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -52.615 uVs 15:03:07 Integrated Voltage Measurement... 15:03:16 Voltage Offset Correction: v_off_ini = 1730.556 uV, v_off_fin = 1735.432 uV 15:03:16 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 75.333 uVs 15:03:16 Integrated Voltage Measurement... 15:03:25 Voltage Offset Correction: v_off_ini = 1727.787 uV, v_off_fin = 1741.747 uV 15:03:25 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -81.106 uVs 15:03:25 Integrated Voltage Measurement... 15:03:34 Voltage Offset Correction: v_off_ini = 1751.291 uV, v_off_fin = 1747.015 uV 15:03:34 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = 57.825 uVs 15:03:34 Integrated Voltage Measurement... 15:03:43 Voltage Offset Correction: v_off_ini = 1760.604 uV, v_off_fin = 1755.331 uV 15:03:43 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -57.850 uVs 15:03:43 Integrated Voltage Measurement... 15:03:52 Voltage Offset Correction: v_off_ini = 1731.984 uV, v_off_fin = 1754.329 uV 15:03:52 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 58.605 uVs 15:03:56 I1X Average and RMS of Measurements VT = -63.889 +- 4.589 uVs I1X = 12.778 +- 0.918 uTm 15:03:59 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 15:04:03 Integrated Voltage Measurement... 15:04:12 Voltage Offset Correction: v_off_ini = 1751.477 uV, v_off_fin = 1745.773 uV 15:04:12 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -175.676 uVs 15:04:12 Integrated Voltage Measurement... 15:04:21 Voltage Offset Correction: v_off_ini = 1722.382 uV, v_off_fin = 1752.141 uV 15:04:21 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 201.213 uVs 15:04:21 Integrated Voltage Measurement... 15:04:30 Voltage Offset Correction: v_off_ini = 1740.366 uV, v_off_fin = 1723.629 uV 15:04:30 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -171.654 uVs 15:04:30 Integrated Voltage Measurement... 15:04:39 Voltage Offset Correction: v_off_ini = 1737.910 uV, v_off_fin = 1746.532 uV 15:04:39 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 206.238 uVs 15:04:39 Integrated Voltage Measurement... 15:04:48 Voltage Offset Correction: v_off_ini = 1727.337 uV, v_off_fin = 1715.638 uV 15:04:48 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -156.886 uVs 15:04:48 Integrated Voltage Measurement... 15:04:57 Voltage Offset Correction: v_off_ini = 1755.971 uV, v_off_fin = 1783.658 uV 15:04:57 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 193.954 uVs 15:05:01 I1Y Average and RMS of Measurements VT = -184.270 +- 6.262 uVs I1Y = -36.852 +- 1.252 uTm 15:05:06 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 15:05:09 Integrated Voltage Measurement... 15:05:18 Voltage Offset Correction: v_off_ini = 1741.125 uV, v_off_fin = 1738.517 uV 15:05:18 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -50.335 uVs 15:05:18 Integrated Voltage Measurement... 15:05:27 Voltage Offset Correction: v_off_ini = 1745.743 uV, v_off_fin = 1770.364 uV 15:05:27 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 41.930 uVs 15:05:27 Integrated Voltage Measurement... 15:05:36 Voltage Offset Correction: v_off_ini = 1745.982 uV, v_off_fin = 1715.088 uV 15:05:36 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -36.788 uVs 15:05:36 Integrated Voltage Measurement... 15:05:45 Voltage Offset Correction: v_off_ini = 1706.646 uV, v_off_fin = 1738.026 uV 15:05:45 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 54.693 uVs 15:05:45 Integrated Voltage Measurement... 15:05:54 Voltage Offset Correction: v_off_ini = 1723.973 uV, v_off_fin = 1715.671 uV 15:05:54 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -24.694 uVs 15:05:54 Integrated Voltage Measurement... 15:06:03 Voltage Offset Correction: v_off_ini = 1719.997 uV, v_off_fin = 1719.873 uV 15:06:03 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 53.202 uVs 15:06:05 I2X Average and RMS of Measurements VT = -43.607 +- 3.298 uVs I2X = 4.012 +- 0.303 uTm^2 15:06:10 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 15:06:13 Integrated Voltage Measurement... 15:06:22 Voltage Offset Correction: v_off_ini = 1714.556 uV, v_off_fin = 1707.335 uV 15:06:22 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -180.651 uVs 15:06:22 Integrated Voltage Measurement... 15:06:31 Voltage Offset Correction: v_off_ini = 1713.954 uV, v_off_fin = 1724.257 uV 15:06:31 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 161.281 uVs 15:06:31 Integrated Voltage Measurement... 15:06:40 Voltage Offset Correction: v_off_ini = 1746.819 uV, v_off_fin = 1713.995 uV 15:06:40 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -188.334 uVs 15:06:40 Integrated Voltage Measurement... 15:06:49 Voltage Offset Correction: v_off_ini = 1723.013 uV, v_off_fin = 1722.503 uV 15:06:49 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 153.553 uVs 15:06:49 Integrated Voltage Measurement... 15:06:58 Voltage Offset Correction: v_off_ini = 1701.735 uV, v_off_fin = 1708.050 uV 15:06:58 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -133.684 uVs 15:06:58 Integrated Voltage Measurement... 15:07:07 Voltage Offset Correction: v_off_ini = 1694.026 uV, v_off_fin = 1708.504 uV 15:07:07 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 187.306 uVs 15:07:10 I2Y Average and RMS of Measurements VT = -167.468 +- 4.931 uVs I2Y = -15.407 +- 0.454 uTm^2 15:07:12 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm