SLAC Magnetic Measurements Date: 09-10-2024 Time: 16:42:59 Project: LCLS-II-HE Magnet Type: Phase Shifter Magnet Name: HE_SXPS_0000 Dataset Number: 1 Run Purpose: CheckRun Run Type: Stretched wire Run Number: 1 Measurement Device: Stretched Wire Operator: YL Comment: Check integrals vs. Gap 16:43:07 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:43:10 Motion Parameters, X Controller, Axis 1 acc = 0.001000 m/s/s dcl = 0.001000 m/s/s vel = 0.001000 m/s pos = 0.000000 m kp = 188.600000 ki = 820.000000 kd = 491.400000 db = 0.000000 16:43:11 Motion Parameters, X Controller, Axis 2 acc = 0.001000 m/s/s dcl = 0.001000 m/s/s vel = 0.001000 m/s pos = 0.000000 m kp = 188.600000 ki = 820.000000 kd = 491.400000 db = 0.000000 16:43:13 Motion Parameters, Y Controller, Axis 1 acc = 0.001000 m/s/s dcl = 0.001000 m/s/s vel = 0.001000 m/s pos = 0.000000 m kp = 188.600000 ki = 820.000000 kd = 491.400000 db = 0.000000 16:43:14 Motion Parameters, Y Controller, Axis 2 acc = 0.001000 m/s/s dcl = 0.001000 m/s/s vel = 0.001000 m/s pos = 0.000000 m kp = 188.600000 ki = 820.000000 kd = 491.400000 db = 0.000000 16:43:25 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:43:30 Integrated Voltage Measurement... 16:43:39 Voltage Offset Correction: v_off_ini = -950.098 uV, v_off_fin = -964.233 uV 16:43:39 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.4997 mm dXY = -0.4998 mm VT = -87.562 uVs 16:43:39 Integrated Voltage Measurement... 16:43:48 Voltage Offset Correction: v_off_ini = -948.582 uV, v_off_fin = -927.200 uV 16:43:48 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 107.778 uVs 16:43:48 Integrated Voltage Measurement... 16:43:57 Voltage Offset Correction: v_off_ini = -910.041 uV, v_off_fin = -926.981 uV 16:43:57 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -101.118 uVs 16:43:57 Integrated Voltage Measurement... 16:44:06 Voltage Offset Correction: v_off_ini = -895.961 uV, v_off_fin = -902.714 uV 16:44:06 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 119.707 uVs 16:44:06 Integrated Voltage Measurement... 16:44:15 Voltage Offset Correction: v_off_ini = -906.870 uV, v_off_fin = -916.309 uV 16:44:15 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.4999 mm dXY = -0.4999 mm VT = -100.412 uVs 16:44:15 Integrated Voltage Measurement... 16:44:24 Voltage Offset Correction: v_off_ini = -910.761 uV, v_off_fin = -928.547 uV 16:44:24 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = 91.755 uVs 16:44:28 I1X Average and RMS of Measurements VT = -101.389 +- 6.413 uVs I1X = 20.280 +- 1.281 uTm 16:44:30 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:44:35 Integrated Voltage Measurement... 16:44:44 Voltage Offset Correction: v_off_ini = -920.365 uV, v_off_fin = -926.070 uV 16:44:44 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -258.074 uVs 16:44:44 Integrated Voltage Measurement... 16:44:53 Voltage Offset Correction: v_off_ini = -915.730 uV, v_off_fin = -919.628 uV 16:44:53 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 223.622 uVs 16:44:53 Integrated Voltage Measurement... 16:45:02 Voltage Offset Correction: v_off_ini = -922.024 uV, v_off_fin = -931.504 uV 16:45:02 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -251.477 uVs 16:45:02 Integrated Voltage Measurement... 16:45:10 Voltage Offset Correction: v_off_ini = -904.088 uV, v_off_fin = -897.082 uV 16:45:11 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 247.910 uVs 16:45:11 Integrated Voltage Measurement... 16:45:19 Voltage Offset Correction: v_off_ini = -910.515 uV, v_off_fin = -921.642 uV 16:45:20 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -248.283 uVs 16:45:20 Integrated Voltage Measurement... 16:45:28 Voltage Offset Correction: v_off_ini = -895.602 uV, v_off_fin = -878.958 uV 16:45:28 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 275.226 uVs 16:45:33 I1Y Average and RMS of Measurements VT = -250.765 +- 8.568 uVs I1Y = -50.150 +- 1.712 uTm 16:45:38 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:45:40 Integrated Voltage Measurement... 16:45:49 Voltage Offset Correction: v_off_ini = -878.354 uV, v_off_fin = -886.757 uV 16:45:49 Integrated Voltage Measurement dXYus = -1.0001 mm, dXYds = 0.0000 mm dXY = -1.0001 mm VT = -117.224 uVs 16:45:49 Integrated Voltage Measurement... 16:45:58 Voltage Offset Correction: v_off_ini = -896.039 uV, v_off_fin = -900.450 uV 16:45:58 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 117.513 uVs 16:45:58 Integrated Voltage Measurement... 16:46:07 Voltage Offset Correction: v_off_ini = -883.857 uV, v_off_fin = -876.329 uV 16:46:07 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -101.624 uVs 16:46:07 Integrated Voltage Measurement... 16:46:16 Voltage Offset Correction: v_off_ini = -872.478 uV, v_off_fin = -875.046 uV 16:46:16 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 105.499 uVs 16:46:16 Integrated Voltage Measurement... 16:46:25 Voltage Offset Correction: v_off_ini = -865.446 uV, v_off_fin = -883.317 uV 16:46:25 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -104.147 uVs 16:46:25 Integrated Voltage Measurement... 16:46:34 Voltage Offset Correction: v_off_ini = -903.472 uV, v_off_fin = -878.257 uV 16:46:34 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 97.377 uVs 16:46:37 I2X Average and RMS of Measurements VT = -107.231 +- 7.259 uVs I2X = 9.866 +- 0.668 uTm^2 16:46:42 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:46:44 Integrated Voltage Measurement... 16:46:53 Voltage Offset Correction: v_off_ini = -883.243 uV, v_off_fin = -888.944 uV 16:46:53 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -265.769 uVs 16:46:53 Integrated Voltage Measurement... 16:47:02 Voltage Offset Correction: v_off_ini = -885.204 uV, v_off_fin = -891.051 uV 16:47:02 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 243.473 uVs 16:47:02 Integrated Voltage Measurement... 16:47:11 Voltage Offset Correction: v_off_ini = -894.145 uV, v_off_fin = -887.030 uV 16:47:11 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -233.143 uVs 16:47:11 Integrated Voltage Measurement... 16:47:20 Voltage Offset Correction: v_off_ini = -885.902 uV, v_off_fin = -900.270 uV 16:47:20 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 241.108 uVs 16:47:20 Integrated Voltage Measurement... 16:47:29 Voltage Offset Correction: v_off_ini = -895.320 uV, v_off_fin = -900.125 uV 16:47:29 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -266.246 uVs 16:47:29 Integrated Voltage Measurement... 16:47:38 Voltage Offset Correction: v_off_ini = -886.889 uV, v_off_fin = -910.107 uV 16:47:38 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 273.068 uVs 16:47:41 I2Y Average and RMS of Measurements VT = -253.801 +- 13.294 uVs I2Y = -23.350 +- 1.223 uTm^2 16:47:50 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:47:55 Integrated Voltage Measurement... 16:48:04 Voltage Offset Correction: v_off_ini = -884.469 uV, v_off_fin = -895.878 uV 16:48:04 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -111.472 uVs 16:48:04 Integrated Voltage Measurement... 16:48:13 Voltage Offset Correction: v_off_ini = -908.219 uV, v_off_fin = -889.138 uV 16:48:13 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 109.692 uVs 16:48:13 Integrated Voltage Measurement... 16:48:22 Voltage Offset Correction: v_off_ini = -910.739 uV, v_off_fin = -911.606 uV 16:48:22 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -83.734 uVs 16:48:22 Integrated Voltage Measurement... 16:48:31 Voltage Offset Correction: v_off_ini = -892.423 uV, v_off_fin = -889.747 uV 16:48:31 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 132.600 uVs 16:48:31 Integrated Voltage Measurement... 16:48:40 Voltage Offset Correction: v_off_ini = -880.462 uV, v_off_fin = -885.176 uV 16:48:40 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -148.116 uVs 16:48:40 Integrated Voltage Measurement... 16:48:49 Voltage Offset Correction: v_off_ini = -899.816 uV, v_off_fin = -898.808 uV 16:48:49 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 138.409 uVs 16:48:53 I1X Average and RMS of Measurements VT = -120.670 +- 16.005 uVs I1X = 24.134 +- 3.201 uTm 16:48:55 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:49:00 Integrated Voltage Measurement... 16:49:09 Voltage Offset Correction: v_off_ini = -889.189 uV, v_off_fin = -910.897 uV 16:49:09 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -227.013 uVs 16:49:09 Integrated Voltage Measurement... 16:49:17 Voltage Offset Correction: v_off_ini = -919.084 uV, v_off_fin = -909.675 uV 16:49:18 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 239.890 uVs 16:49:18 Integrated Voltage Measurement... 16:49:26 Voltage Offset Correction: v_off_ini = -903.492 uV, v_off_fin = -906.797 uV 16:49:26 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -235.318 uVs 16:49:27 Integrated Voltage Measurement... 16:49:35 Voltage Offset Correction: v_off_ini = -922.065 uV, v_off_fin = -900.697 uV 16:49:35 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 266.578 uVs 16:49:35 Integrated Voltage Measurement... 16:49:44 Voltage Offset Correction: v_off_ini = -920.454 uV, v_off_fin = -896.067 uV 16:49:44 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -231.162 uVs 16:49:44 Integrated Voltage Measurement... 16:49:53 Voltage Offset Correction: v_off_ini = -928.473 uV, v_off_fin = -904.976 uV 16:49:53 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 247.444 uVs 16:49:58 I1Y Average and RMS of Measurements VT = -241.234 +- 7.272 uVs I1Y = -48.242 +- 1.454 uTm 16:50:03 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:50:05 Integrated Voltage Measurement... 16:50:14 Voltage Offset Correction: v_off_ini = -912.699 uV, v_off_fin = -922.655 uV 16:50:14 Integrated Voltage Measurement dXYus = -1.0001 mm, dXYds = 0.0000 mm dXY = -1.0001 mm VT = -121.069 uVs 16:50:14 Integrated Voltage Measurement... 16:50:23 Voltage Offset Correction: v_off_ini = -922.105 uV, v_off_fin = -950.715 uV 16:50:23 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 73.236 uVs 16:50:23 Integrated Voltage Measurement... 16:50:32 Voltage Offset Correction: v_off_ini = -927.945 uV, v_off_fin = -915.411 uV 16:50:32 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -89.182 uVs 16:50:32 Integrated Voltage Measurement... 16:50:41 Voltage Offset Correction: v_off_ini = -919.998 uV, v_off_fin = -913.468 uV 16:50:41 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 95.122 uVs 16:50:41 Integrated Voltage Measurement... 16:50:50 Voltage Offset Correction: v_off_ini = -909.533 uV, v_off_fin = -916.213 uV 16:50:50 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -140.220 uVs 16:50:50 Integrated Voltage Measurement... 16:50:59 Voltage Offset Correction: v_off_ini = -909.509 uV, v_off_fin = -912.619 uV 16:50:59 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 135.359 uVs 16:51:02 I2X Average and RMS of Measurements VT = -109.032 +- 20.437 uVs I2X = 10.031 +- 1.881 uTm^2 16:51:07 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:51:09 Integrated Voltage Measurement... 16:51:18 Voltage Offset Correction: v_off_ini = -926.787 uV, v_off_fin = -905.934 uV 16:51:18 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -254.390 uVs 16:51:18 Integrated Voltage Measurement... 16:51:27 Voltage Offset Correction: v_off_ini = -921.692 uV, v_off_fin = -938.116 uV 16:51:27 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 222.866 uVs 16:51:27 Integrated Voltage Measurement... 16:51:36 Voltage Offset Correction: v_off_ini = -930.502 uV, v_off_fin = -934.366 uV 16:51:36 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -248.071 uVs 16:51:36 Integrated Voltage Measurement... 16:51:45 Voltage Offset Correction: v_off_ini = -956.007 uV, v_off_fin = -950.540 uV 16:51:45 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 286.187 uVs 16:51:45 Integrated Voltage Measurement... 16:51:54 Voltage Offset Correction: v_off_ini = -962.566 uV, v_off_fin = -954.894 uV 16:51:54 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -250.111 uVs 16:51:54 Integrated Voltage Measurement... 16:52:03 Voltage Offset Correction: v_off_ini = -915.712 uV, v_off_fin = -928.557 uV 16:52:03 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 257.177 uVs 16:52:06 I2Y Average and RMS of Measurements VT = -253.134 +- 11.641 uVs I2Y = -23.288 +- 1.071 uTm^2 16:52:15 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:52:19 Integrated Voltage Measurement... 16:52:28 Voltage Offset Correction: v_off_ini = -928.261 uV, v_off_fin = -935.035 uV 16:52:28 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -106.357 uVs 16:52:28 Integrated Voltage Measurement... 16:52:37 Voltage Offset Correction: v_off_ini = -955.143 uV, v_off_fin = -925.820 uV 16:52:37 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 120.449 uVs 16:52:37 Integrated Voltage Measurement... 16:52:46 Voltage Offset Correction: v_off_ini = -923.048 uV, v_off_fin = -938.165 uV 16:52:46 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -107.421 uVs 16:52:46 Integrated Voltage Measurement... 16:52:55 Voltage Offset Correction: v_off_ini = -941.076 uV, v_off_fin = -937.997 uV 16:52:55 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 91.001 uVs 16:52:55 Integrated Voltage Measurement... 16:53:04 Voltage Offset Correction: v_off_ini = -927.941 uV, v_off_fin = -936.134 uV 16:53:04 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -110.795 uVs 16:53:04 Integrated Voltage Measurement... 16:53:13 Voltage Offset Correction: v_off_ini = -934.112 uV, v_off_fin = -931.312 uV 16:53:13 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 111.999 uVs 16:53:17 I1X Average and RMS of Measurements VT = -108.004 +- 6.271 uVs I1X = 21.601 +- 1.254 uTm 16:53:20 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:53:24 Integrated Voltage Measurement... 16:53:33 Voltage Offset Correction: v_off_ini = -939.247 uV, v_off_fin = -952.137 uV 16:53:33 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -243.677 uVs 16:53:33 Integrated Voltage Measurement... 16:53:42 Voltage Offset Correction: v_off_ini = -961.325 uV, v_off_fin = -963.571 uV 16:53:42 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 207.982 uVs 16:53:42 Integrated Voltage Measurement... 16:53:51 Voltage Offset Correction: v_off_ini = -964.934 uV, v_off_fin = -988.491 uV 16:53:51 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -213.232 uVs 16:53:51 Integrated Voltage Measurement... 16:54:00 Voltage Offset Correction: v_off_ini = -987.200 uV, v_off_fin = -951.879 uV 16:54:00 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 219.430 uVs 16:54:00 Integrated Voltage Measurement... 16:54:09 Voltage Offset Correction: v_off_ini = -953.051 uV, v_off_fin = -984.096 uV 16:54:09 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -234.041 uVs 16:54:09 Integrated Voltage Measurement... 16:54:18 Voltage Offset Correction: v_off_ini = -999.675 uV, v_off_fin = -955.910 uV 16:54:18 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 234.769 uVs 16:54:22 I1Y Average and RMS of Measurements VT = -225.522 +- 7.382 uVs I1Y = -45.101 +- 1.476 uTm 16:54:27 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:54:30 Integrated Voltage Measurement... 16:54:39 Voltage Offset Correction: v_off_ini = -993.704 uV, v_off_fin = -978.011 uV 16:54:39 Integrated Voltage Measurement dXYus = -1.0002 mm, dXYds = 0.0000 mm dXY = -1.0002 mm VT = -101.824 uVs 16:54:39 Integrated Voltage Measurement... 16:54:48 Voltage Offset Correction: v_off_ini = -968.261 uV, v_off_fin = -956.527 uV 16:54:48 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 100.008 uVs 16:54:48 Integrated Voltage Measurement... 16:54:57 Voltage Offset Correction: v_off_ini = -984.896 uV, v_off_fin = -970.645 uV 16:54:57 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -100.679 uVs 16:54:57 Integrated Voltage Measurement... 16:55:06 Voltage Offset Correction: v_off_ini = -1002.703 uV, v_off_fin = -975.617 uV 16:55:06 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 100.577 uVs 16:55:06 Integrated Voltage Measurement... 16:55:15 Voltage Offset Correction: v_off_ini = -973.435 uV, v_off_fin = -983.230 uV 16:55:15 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -82.482 uVs 16:55:15 Integrated Voltage Measurement... 16:55:24 Voltage Offset Correction: v_off_ini = -980.759 uV, v_off_fin = -987.060 uV 16:55:24 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 111.986 uVs 16:55:26 I2X Average and RMS of Measurements VT = -99.593 +- 1.672 uVs I2X = 9.162 +- 0.153 uTm^2 16:55:31 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:55:34 Integrated Voltage Measurement... 16:55:43 Voltage Offset Correction: v_off_ini = -978.443 uV, v_off_fin = -983.558 uV 16:55:43 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -223.798 uVs 16:55:43 Integrated Voltage Measurement... 16:55:52 Voltage Offset Correction: v_off_ini = -1014.662 uV, v_off_fin = -989.572 uV 16:55:52 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 260.421 uVs 16:55:52 Integrated Voltage Measurement... 16:56:01 Voltage Offset Correction: v_off_ini = -998.425 uV, v_off_fin = -989.828 uV 16:56:01 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -230.025 uVs 16:56:01 Integrated Voltage Measurement... 16:56:10 Voltage Offset Correction: v_off_ini = -992.166 uV, v_off_fin = -1002.871 uV 16:56:10 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 239.322 uVs 16:56:10 Integrated Voltage Measurement... 16:56:19 Voltage Offset Correction: v_off_ini = -1013.736 uV, v_off_fin = -1001.154 uV 16:56:19 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -207.895 uVs 16:56:19 Integrated Voltage Measurement... 16:56:28 Voltage Offset Correction: v_off_ini = -986.445 uV, v_off_fin = -974.739 uV 16:56:28 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 248.238 uVs 16:56:30 I2Y Average and RMS of Measurements VT = -234.950 +- 5.736 uVs I2Y = -21.615 +- 0.528 uTm^2 16:56:40 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:56:44 Integrated Voltage Measurement... 16:56:53 Voltage Offset Correction: v_off_ini = -980.201 uV, v_off_fin = -966.175 uV 16:56:53 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.4998 mm dXY = -0.4999 mm VT = -97.614 uVs 16:56:53 Integrated Voltage Measurement... 16:57:02 Voltage Offset Correction: v_off_ini = -1000.116 uV, v_off_fin = -990.037 uV 16:57:02 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 95.861 uVs 16:57:02 Integrated Voltage Measurement... 16:57:11 Voltage Offset Correction: v_off_ini = -1001.901 uV, v_off_fin = -972.243 uV 16:57:11 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -95.279 uVs 16:57:11 Integrated Voltage Measurement... 16:57:20 Voltage Offset Correction: v_off_ini = -970.099 uV, v_off_fin = -971.502 uV 16:57:20 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 99.298 uVs 16:57:20 Integrated Voltage Measurement... 16:57:29 Voltage Offset Correction: v_off_ini = -982.085 uV, v_off_fin = -972.986 uV 16:57:29 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -118.893 uVs 16:57:29 Integrated Voltage Measurement... 16:57:38 Voltage Offset Correction: v_off_ini = -974.820 uV, v_off_fin = -953.881 uV 16:57:38 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 107.885 uVs 16:57:42 I1X Average and RMS of Measurements VT = -102.472 +- 7.723 uVs I1X = 20.496 +- 1.544 uTm 16:57:45 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:57:49 Integrated Voltage Measurement... 16:57:58 Voltage Offset Correction: v_off_ini = -977.325 uV, v_off_fin = -969.981 uV 16:57:58 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -190.228 uVs 16:57:58 Integrated Voltage Measurement... 16:58:07 Voltage Offset Correction: v_off_ini = -1012.994 uV, v_off_fin = -995.059 uV 16:58:07 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 253.897 uVs 16:58:07 Integrated Voltage Measurement... 16:58:16 Voltage Offset Correction: v_off_ini = -1004.789 uV, v_off_fin = -997.876 uV 16:58:16 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -213.928 uVs 16:58:16 Integrated Voltage Measurement... 16:58:25 Voltage Offset Correction: v_off_ini = -1001.086 uV, v_off_fin = -967.080 uV 16:58:25 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 226.997 uVs 16:58:25 Integrated Voltage Measurement... 16:58:34 Voltage Offset Correction: v_off_ini = -988.556 uV, v_off_fin = -1007.070 uV 16:58:34 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -213.678 uVs 16:58:34 Integrated Voltage Measurement... 16:58:43 Voltage Offset Correction: v_off_ini = -1001.721 uV, v_off_fin = -962.840 uV 16:58:43 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 224.686 uVs 16:58:47 I1Y Average and RMS of Measurements VT = -220.569 +- 1.178 uVs I1Y = -44.109 +- 0.236 uTm 16:58:52 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:58:55 Integrated Voltage Measurement... 16:59:04 Voltage Offset Correction: v_off_ini = -992.570 uV, v_off_fin = -973.858 uV 16:59:04 Integrated Voltage Measurement dXYus = -1.0002 mm, dXYds = 0.0000 mm dXY = -1.0002 mm VT = -100.980 uVs 16:59:04 Integrated Voltage Measurement... 16:59:13 Voltage Offset Correction: v_off_ini = -981.312 uV, v_off_fin = -970.277 uV 16:59:13 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 69.732 uVs 16:59:13 Integrated Voltage Measurement... 16:59:22 Voltage Offset Correction: v_off_ini = -975.968 uV, v_off_fin = -976.059 uV 16:59:22 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -101.948 uVs 16:59:22 Integrated Voltage Measurement... 16:59:31 Voltage Offset Correction: v_off_ini = -966.819 uV, v_off_fin = -994.792 uV 16:59:31 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 117.526 uVs 16:59:31 Integrated Voltage Measurement... 16:59:40 Voltage Offset Correction: v_off_ini = -970.798 uV, v_off_fin = -964.534 uV 16:59:40 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -105.443 uVs 16:59:40 Integrated Voltage Measurement... 16:59:49 Voltage Offset Correction: v_off_ini = -971.975 uV, v_off_fin = -968.128 uV 16:59:49 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 126.835 uVs 16:59:51 I2X Average and RMS of Measurements VT = -103.744 +- 13.262 uVs I2X = 9.544 +- 1.221 uTm^2 16:59:56 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:59:59 Integrated Voltage Measurement... 17:00:08 Voltage Offset Correction: v_off_ini = -965.296 uV, v_off_fin = -950.298 uV 17:00:08 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -222.488 uVs 17:00:08 Integrated Voltage Measurement... 17:00:17 Voltage Offset Correction: v_off_ini = -966.252 uV, v_off_fin = -973.766 uV 17:00:17 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 195.785 uVs 17:00:17 Integrated Voltage Measurement... 17:00:26 Voltage Offset Correction: v_off_ini = -1003.418 uV, v_off_fin = -1010.673 uV 17:00:26 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = -221.438 uVs 17:00:26 Integrated Voltage Measurement... 17:00:35 Voltage Offset Correction: v_off_ini = -982.963 uV, v_off_fin = -981.151 uV 17:00:35 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 234.404 uVs 17:00:35 Integrated Voltage Measurement... 17:00:44 Voltage Offset Correction: v_off_ini = -1014.643 uV, v_off_fin = -992.687 uV 17:00:44 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -212.324 uVs 17:00:44 Integrated Voltage Measurement... 17:00:53 Voltage Offset Correction: v_off_ini = -982.010 uV, v_off_fin = -983.593 uV 17:00:53 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 227.899 uVs 17:00:55 I2Y Average and RMS of Measurements VT = -219.057 +- 7.705 uVs I2Y = -20.154 +- 0.709 uTm^2 17:01:05 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 17:01:09 Integrated Voltage Measurement... 17:01:18 Voltage Offset Correction: v_off_ini = -983.032 uV, v_off_fin = -984.697 uV 17:01:18 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -144.299 uVs 17:01:18 Integrated Voltage Measurement... 17:01:27 Voltage Offset Correction: v_off_ini = -1011.036 uV, v_off_fin = -1020.890 uV 17:01:27 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 121.159 uVs 17:01:27 Integrated Voltage Measurement... 17:01:36 Voltage Offset Correction: v_off_ini = -996.950 uV, v_off_fin = -988.723 uV 17:01:36 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -119.570 uVs 17:01:36 Integrated Voltage Measurement... 17:01:45 Voltage Offset Correction: v_off_ini = -989.624 uV, v_off_fin = -972.117 uV 17:01:45 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 103.328 uVs 17:01:45 Integrated Voltage Measurement... 17:01:54 Voltage Offset Correction: v_off_ini = -989.255 uV, v_off_fin = -966.856 uV 17:01:54 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -135.597 uVs 17:01:54 Integrated Voltage Measurement... 17:02:03 Voltage Offset Correction: v_off_ini = -970.063 uV, v_off_fin = -953.958 uV 17:02:03 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = 109.133 uVs 17:02:07 I1X Average and RMS of Measurements VT = -122.181 +- 8.688 uVs I1X = 24.437 +- 1.738 uTm 17:02:09 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 17:02:14 Integrated Voltage Measurement... 17:02:23 Voltage Offset Correction: v_off_ini = -950.042 uV, v_off_fin = -975.457 uV 17:02:23 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -210.210 uVs 17:02:23 Integrated Voltage Measurement... 17:02:32 Voltage Offset Correction: v_off_ini = -1007.565 uV, v_off_fin = -987.894 uV 17:02:32 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 238.344 uVs 17:02:32 Integrated Voltage Measurement... 17:02:41 Voltage Offset Correction: v_off_ini = -982.940 uV, v_off_fin = -978.631 uV 17:02:41 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -193.124 uVs 17:02:41 Integrated Voltage Measurement... 17:02:50 Voltage Offset Correction: v_off_ini = -951.437 uV, v_off_fin = -945.120 uV 17:02:50 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 235.251 uVs 17:02:50 Integrated Voltage Measurement... 17:02:59 Voltage Offset Correction: v_off_ini = -950.415 uV, v_off_fin = -960.072 uV 17:02:59 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -204.523 uVs 17:02:59 Integrated Voltage Measurement... 17:03:07 Voltage Offset Correction: v_off_ini = -962.860 uV, v_off_fin = -949.671 uV 17:03:08 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 191.569 uVs 17:03:12 I1Y Average and RMS of Measurements VT = -212.170 +- 10.803 uVs I1Y = -42.430 +- 2.161 uTm 17:03:17 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 17:03:19 Integrated Voltage Measurement... 17:03:28 Voltage Offset Correction: v_off_ini = -973.747 uV, v_off_fin = -940.565 uV 17:03:28 Integrated Voltage Measurement dXYus = -1.0002 mm, dXYds = 0.0000 mm dXY = -1.0002 mm VT = -82.600 uVs 17:03:28 Integrated Voltage Measurement... 17:03:37 Voltage Offset Correction: v_off_ini = -930.056 uV, v_off_fin = -949.579 uV 17:03:37 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 107.438 uVs 17:03:37 Integrated Voltage Measurement... 17:03:46 Voltage Offset Correction: v_off_ini = -934.980 uV, v_off_fin = -953.126 uV 17:03:46 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -127.328 uVs 17:03:46 Integrated Voltage Measurement... 17:03:55 Voltage Offset Correction: v_off_ini = -956.321 uV, v_off_fin = -951.817 uV 17:03:55 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 128.480 uVs 17:03:55 Integrated Voltage Measurement... 17:04:04 Voltage Offset Correction: v_off_ini = -946.864 uV, v_off_fin = -960.148 uV 17:04:04 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -102.072 uVs 17:04:04 Integrated Voltage Measurement... 17:04:13 Voltage Offset Correction: v_off_ini = -951.433 uV, v_off_fin = -944.448 uV 17:04:13 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 123.175 uVs 17:04:16 I2X Average and RMS of Measurements VT = -111.849 +- 13.437 uVs I2X = 10.290 +- 1.237 uTm^2 17:04:21 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 17:04:23 Integrated Voltage Measurement... 17:04:32 Voltage Offset Correction: v_off_ini = -966.274 uV, v_off_fin = -956.374 uV 17:04:32 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -221.812 uVs 17:04:32 Integrated Voltage Measurement... 17:04:41 Voltage Offset Correction: v_off_ini = -957.361 uV, v_off_fin = -953.501 uV 17:04:41 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 213.300 uVs 17:04:41 Integrated Voltage Measurement... 17:04:50 Voltage Offset Correction: v_off_ini = -970.799 uV, v_off_fin = -985.200 uV 17:04:50 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -207.880 uVs 17:04:50 Integrated Voltage Measurement... 17:04:59 Voltage Offset Correction: v_off_ini = -971.561 uV, v_off_fin = -971.219 uV 17:04:59 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 218.935 uVs 17:04:59 Integrated Voltage Measurement... 17:05:08 Voltage Offset Correction: v_off_ini = -963.056 uV, v_off_fin = -983.096 uV 17:05:08 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -196.699 uVs 17:05:08 Integrated Voltage Measurement... 17:05:17 Voltage Offset Correction: v_off_ini = -979.002 uV, v_off_fin = -973.228 uV 17:05:17 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 232.138 uVs 17:05:20 I2Y Average and RMS of Measurements VT = -215.127 +- 1.766 uVs I2Y = -19.792 +- 0.162 uTm^2 17:05:30 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 17:05:34 Integrated Voltage Measurement... 17:05:43 Voltage Offset Correction: v_off_ini = -970.337 uV, v_off_fin = -997.380 uV 17:05:43 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4998 mm dXY = -0.4999 mm VT = -138.909 uVs 17:05:43 Integrated Voltage Measurement... 17:05:52 Voltage Offset Correction: v_off_ini = -980.063 uV, v_off_fin = -984.006 uV 17:05:52 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 108.742 uVs 17:05:52 Integrated Voltage Measurement... 17:06:01 Voltage Offset Correction: v_off_ini = -995.696 uV, v_off_fin = -997.592 uV 17:06:01 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -105.662 uVs 17:06:01 Integrated Voltage Measurement... 17:06:10 Voltage Offset Correction: v_off_ini = -983.806 uV, v_off_fin = -996.954 uV 17:06:10 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = 111.496 uVs 17:06:10 Integrated Voltage Measurement... 17:06:19 Voltage Offset Correction: v_off_ini = -991.465 uV, v_off_fin = -965.746 uV 17:06:19 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -102.529 uVs 17:06:19 Integrated Voltage Measurement... 17:06:28 Voltage Offset Correction: v_off_ini = -990.644 uV, v_off_fin = -988.193 uV 17:06:28 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 138.393 uVs 17:06:32 I1X Average and RMS of Measurements VT = -117.622 +- 6.540 uVs I1X = 23.526 +- 1.308 uTm 17:06:34 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 17:06:39 Integrated Voltage Measurement... 17:06:48 Voltage Offset Correction: v_off_ini = -987.334 uV, v_off_fin = -976.355 uV 17:06:48 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -204.437 uVs 17:06:48 Integrated Voltage Measurement... 17:06:57 Voltage Offset Correction: v_off_ini = -997.769 uV, v_off_fin = -987.201 uV 17:06:57 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 206.656 uVs 17:06:57 Integrated Voltage Measurement... 17:07:05 Voltage Offset Correction: v_off_ini = -975.216 uV, v_off_fin = -1011.537 uV 17:07:06 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -210.526 uVs 17:07:06 Integrated Voltage Measurement... 17:07:14 Voltage Offset Correction: v_off_ini = -980.724 uV, v_off_fin = -990.527 uV 17:07:14 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 228.127 uVs 17:07:14 Integrated Voltage Measurement... 17:07:23 Voltage Offset Correction: v_off_ini = -1014.901 uV, v_off_fin = -1006.945 uV 17:07:23 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -185.035 uVs 17:07:23 Integrated Voltage Measurement... 17:07:32 Voltage Offset Correction: v_off_ini = -953.374 uV, v_off_fin = -969.569 uV 17:07:32 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 192.317 uVs 17:07:37 I1Y Average and RMS of Measurements VT = -204.516 +- 12.534 uVs I1Y = -40.901 +- 2.506 uTm 17:07:42 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 17:07:44 Integrated Voltage Measurement... 17:07:53 Voltage Offset Correction: v_off_ini = -974.663 uV, v_off_fin = -986.138 uV 17:07:53 Integrated Voltage Measurement dXYus = -1.0002 mm, dXYds = 0.0000 mm dXY = -1.0002 mm VT = -110.235 uVs 17:07:53 Integrated Voltage Measurement... 17:08:02 Voltage Offset Correction: v_off_ini = -964.714 uV, v_off_fin = -984.270 uV 17:08:02 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 88.010 uVs 17:08:02 Integrated Voltage Measurement... 17:08:11 Voltage Offset Correction: v_off_ini = -1000.099 uV, v_off_fin = -994.764 uV 17:08:11 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -93.598 uVs 17:08:11 Integrated Voltage Measurement... 17:08:20 Voltage Offset Correction: v_off_ini = -995.119 uV, v_off_fin = -976.740 uV 17:08:20 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 123.561 uVs 17:08:20 Integrated Voltage Measurement... 17:08:29 Voltage Offset Correction: v_off_ini = -1013.336 uV, v_off_fin = -989.907 uV 17:08:29 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -103.965 uVs 17:08:29 Integrated Voltage Measurement... 17:08:38 Voltage Offset Correction: v_off_ini = -1007.318 uV, v_off_fin = -973.125 uV 17:08:38 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 117.230 uVs 17:08:41 I2X Average and RMS of Measurements VT = -106.100 +- 5.002 uVs I2X = 9.761 +- 0.461 uTm^2 17:08:46 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 17:08:48 Integrated Voltage Measurement... 17:08:57 Voltage Offset Correction: v_off_ini = -990.675 uV, v_off_fin = -1011.107 uV 17:08:57 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -243.282 uVs 17:08:57 Integrated Voltage Measurement... 17:09:06 Voltage Offset Correction: v_off_ini = -1026.426 uV, v_off_fin = -1024.360 uV 17:09:06 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = -0.0001 mm dXY = -1.0000 mm VT = 192.818 uVs 17:09:06 Integrated Voltage Measurement... 17:09:15 Voltage Offset Correction: v_off_ini = -1011.155 uV, v_off_fin = -1012.163 uV 17:09:15 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -213.931 uVs 17:09:15 Integrated Voltage Measurement... 17:09:24 Voltage Offset Correction: v_off_ini = -991.302 uV, v_off_fin = -993.765 uV 17:09:24 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0001 mm dXY = -1.0000 mm VT = 199.017 uVs 17:09:24 Integrated Voltage Measurement... 17:09:33 Voltage Offset Correction: v_off_ini = -1004.419 uV, v_off_fin = -985.883 uV 17:09:33 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -215.763 uVs 17:09:33 Integrated Voltage Measurement... 17:09:42 Voltage Offset Correction: v_off_ini = -1009.126 uV, v_off_fin = -985.240 uV 17:09:42 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 206.876 uVs 17:09:45 I2Y Average and RMS of Measurements VT = -211.948 +- 4.747 uVs I2Y = -19.499 +- 0.437 uTm^2 17:09:54 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 17:09:59 Integrated Voltage Measurement... 17:10:08 Voltage Offset Correction: v_off_ini = -1012.036 uV, v_off_fin = -1015.948 uV 17:10:08 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -120.122 uVs 17:10:08 Integrated Voltage Measurement... 17:10:17 Voltage Offset Correction: v_off_ini = -1014.785 uV, v_off_fin = -1007.080 uV 17:10:17 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 87.765 uVs 17:10:17 Integrated Voltage Measurement... 17:10:26 Voltage Offset Correction: v_off_ini = -977.397 uV, v_off_fin = -1002.695 uV 17:10:26 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -124.053 uVs 17:10:26 Integrated Voltage Measurement... 17:10:34 Voltage Offset Correction: v_off_ini = -1011.080 uV, v_off_fin = -1005.952 uV 17:10:35 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 137.847 uVs 17:10:35 Integrated Voltage Measurement... 17:10:43 Voltage Offset Correction: v_off_ini = -994.997 uV, v_off_fin = -1007.961 uV 17:10:44 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -122.894 uVs 17:10:44 Integrated Voltage Measurement... 17:10:52 Voltage Offset Correction: v_off_ini = -1001.519 uV, v_off_fin = -1016.818 uV 17:10:53 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 92.149 uVs 17:10:57 I1X Average and RMS of Measurements VT = -114.138 +- 11.977 uVs I1X = 22.828 +- 2.395 uTm 17:10:59 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 17:11:03 Integrated Voltage Measurement... 17:11:12 Voltage Offset Correction: v_off_ini = -1017.009 uV, v_off_fin = -1012.844 uV 17:11:12 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -178.418 uVs 17:11:12 Integrated Voltage Measurement... 17:11:21 Voltage Offset Correction: v_off_ini = -988.473 uV, v_off_fin = -999.757 uV 17:11:21 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5002 mm dXY = -0.5001 mm VT = 176.047 uVs 17:11:21 Integrated Voltage Measurement... 17:11:30 Voltage Offset Correction: v_off_ini = -994.499 uV, v_off_fin = -1000.305 uV 17:11:30 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.5002 mm dXY = 0.5000 mm VT = -193.303 uVs 17:11:30 Integrated Voltage Measurement... 17:11:39 Voltage Offset Correction: v_off_ini = -1027.972 uV, v_off_fin = -1003.586 uV 17:11:39 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5002 mm dXY = -0.5001 mm VT = 200.804 uVs 17:11:39 Integrated Voltage Measurement... 17:11:48 Voltage Offset Correction: v_off_ini = -1004.906 uV, v_off_fin = -990.652 uV 17:11:48 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.5002 mm dXY = 0.5000 mm VT = -202.045 uVs 17:11:48 Integrated Voltage Measurement... 17:11:57 Voltage Offset Correction: v_off_ini = -978.000 uV, v_off_fin = -1004.554 uV 17:11:57 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5002 mm dXY = -0.5001 mm VT = 180.867 uVs 17:12:01 I1Y Average and RMS of Measurements VT = -188.581 +- 8.343 uVs I1Y = -37.711 +- 1.668 uTm 17:12:06 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 17:12:09 Integrated Voltage Measurement... 17:12:18 Voltage Offset Correction: v_off_ini = -996.018 uV, v_off_fin = -1003.016 uV 17:12:18 Integrated Voltage Measurement dXYus = -1.0002 mm, dXYds = 0.0000 mm dXY = -1.0002 mm VT = -83.715 uVs 17:12:18 Integrated Voltage Measurement... 17:12:27 Voltage Offset Correction: v_off_ini = -972.256 uV, v_off_fin = -981.162 uV 17:12:27 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 109.404 uVs 17:12:27 Integrated Voltage Measurement... 17:12:36 Voltage Offset Correction: v_off_ini = -979.447 uV, v_off_fin = -980.211 uV 17:12:36 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -103.275 uVs 17:12:36 Integrated Voltage Measurement... 17:12:45 Voltage Offset Correction: v_off_ini = -975.975 uV, v_off_fin = -976.727 uV 17:12:45 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 87.611 uVs 17:12:45 Integrated Voltage Measurement... 17:12:54 Voltage Offset Correction: v_off_ini = -982.694 uV, v_off_fin = -958.450 uV 17:12:54 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -112.588 uVs 17:12:54 Integrated Voltage Measurement... 17:13:03 Voltage Offset Correction: v_off_ini = -959.706 uV, v_off_fin = -968.488 uV 17:13:03 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 82.960 uVs 17:13:05 I2X Average and RMS of Measurements VT = -96.592 +- 0.952 uVs I2X = 8.886 +- 0.088 uTm^2 17:13:11 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 17:13:13 Integrated Voltage Measurement... 17:13:22 Voltage Offset Correction: v_off_ini = -971.848 uV, v_off_fin = -972.149 uV 17:13:22 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -189.716 uVs 17:13:22 Integrated Voltage Measurement... 17:13:31 Voltage Offset Correction: v_off_ini = -962.784 uV, v_off_fin = -960.321 uV 17:13:31 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 180.655 uVs 17:13:31 Integrated Voltage Measurement... 17:13:40 Voltage Offset Correction: v_off_ini = -949.791 uV, v_off_fin = -959.321 uV 17:13:40 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -196.577 uVs 17:13:40 Integrated Voltage Measurement... 17:13:49 Voltage Offset Correction: v_off_ini = -956.189 uV, v_off_fin = -955.063 uV 17:13:49 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 208.141 uVs 17:13:49 Integrated Voltage Measurement... 17:13:58 Voltage Offset Correction: v_off_ini = -948.982 uV, v_off_fin = -958.099 uV 17:13:58 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -166.420 uVs 17:13:58 Integrated Voltage Measurement... 17:14:07 Voltage Offset Correction: v_off_ini = -941.865 uV, v_off_fin = -938.832 uV 17:14:07 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 230.306 uVs 17:14:10 I2Y Average and RMS of Measurements VT = -195.302 +- 7.337 uVs I2Y = -17.968 +- 0.675 uTm^2 17:14:19 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 17:14:23 Integrated Voltage Measurement... 17:14:32 Voltage Offset Correction: v_off_ini = -952.003 uV, v_off_fin = -937.036 uV 17:14:32 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -113.681 uVs 17:14:32 Integrated Voltage Measurement... 17:14:41 Voltage Offset Correction: v_off_ini = -921.629 uV, v_off_fin = -954.247 uV 17:14:41 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 114.938 uVs 17:14:41 Integrated Voltage Measurement... 17:14:50 Voltage Offset Correction: v_off_ini = -963.185 uV, v_off_fin = -950.249 uV 17:14:50 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -121.341 uVs 17:14:50 Integrated Voltage Measurement... 17:14:59 Voltage Offset Correction: v_off_ini = -952.558 uV, v_off_fin = -931.173 uV 17:14:59 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.4999 mm dXY = 0.4999 mm VT = 118.683 uVs 17:14:59 Integrated Voltage Measurement... 17:15:08 Voltage Offset Correction: v_off_ini = -944.443 uV, v_off_fin = -965.550 uV 17:15:08 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -89.591 uVs 17:15:08 Integrated Voltage Measurement... 17:15:17 Voltage Offset Correction: v_off_ini = -963.097 uV, v_off_fin = -962.423 uV 17:15:17 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 108.192 uVs 17:15:22 I1X Average and RMS of Measurements VT = -111.071 +- 8.921 uVs I1X = 22.216 +- 1.785 uTm 17:15:24 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 17:15:28 Integrated Voltage Measurement... 17:15:37 Voltage Offset Correction: v_off_ini = -964.022 uV, v_off_fin = -1008.265 uV 17:15:37 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -180.639 uVs 17:15:37 Integrated Voltage Measurement... 17:15:46 Voltage Offset Correction: v_off_ini = -957.635 uV, v_off_fin = -960.444 uV 17:15:46 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 187.426 uVs 17:15:46 Integrated Voltage Measurement... 17:15:55 Voltage Offset Correction: v_off_ini = -966.126 uV, v_off_fin = -971.969 uV 17:15:55 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -178.319 uVs 17:15:55 Integrated Voltage Measurement... 17:16:04 Voltage Offset Correction: v_off_ini = -975.140 uV, v_off_fin = -980.151 uV 17:16:04 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 183.222 uVs 17:16:04 Integrated Voltage Measurement... 17:16:13 Voltage Offset Correction: v_off_ini = -980.957 uV, v_off_fin = -988.879 uV 17:16:13 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -203.940 uVs 17:16:13 Integrated Voltage Measurement... 17:16:22 Voltage Offset Correction: v_off_ini = -965.314 uV, v_off_fin = -979.470 uV 17:16:22 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 179.945 uVs 17:16:26 I1Y Average and RMS of Measurements VT = -185.582 +- 4.691 uVs I1Y = -37.115 +- 0.939 uTm 17:16:31 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 17:16:34 Integrated Voltage Measurement... 17:16:43 Voltage Offset Correction: v_off_ini = -1004.018 uV, v_off_fin = -988.197 uV 17:16:43 Integrated Voltage Measurement dXYus = -1.0002 mm, dXYds = 0.0000 mm dXY = -1.0002 mm VT = -82.018 uVs 17:16:43 Integrated Voltage Measurement... 17:16:52 Voltage Offset Correction: v_off_ini = -976.501 uV, v_off_fin = -967.819 uV 17:16:52 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 87.016 uVs 17:16:52 Integrated Voltage Measurement... 17:17:01 Voltage Offset Correction: v_off_ini = -969.800 uV, v_off_fin = -982.328 uV 17:17:01 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -99.375 uVs 17:17:01 Integrated Voltage Measurement... 17:17:10 Voltage Offset Correction: v_off_ini = -983.492 uV, v_off_fin = -974.767 uV 17:17:10 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 94.313 uVs 17:17:10 Integrated Voltage Measurement... 17:17:19 Voltage Offset Correction: v_off_ini = -970.978 uV, v_off_fin = -978.479 uV 17:17:19 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -97.333 uVs 17:17:19 Integrated Voltage Measurement... 17:17:28 Voltage Offset Correction: v_off_ini = -972.905 uV, v_off_fin = -970.801 uV 17:17:28 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 94.819 uVs 17:17:30 I2X Average and RMS of Measurements VT = -92.479 +- 5.639 uVs I2X = 8.508 +- 0.519 uTm^2 17:17:35 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 17:17:38 Integrated Voltage Measurement... 17:17:47 Voltage Offset Correction: v_off_ini = -961.388 uV, v_off_fin = -983.339 uV 17:17:47 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -202.712 uVs 17:17:47 Integrated Voltage Measurement... 17:17:56 Voltage Offset Correction: v_off_ini = -969.475 uV, v_off_fin = -980.597 uV 17:17:56 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 192.757 uVs 17:17:56 Integrated Voltage Measurement... 17:18:05 Voltage Offset Correction: v_off_ini = -1012.849 uV, v_off_fin = -999.088 uV 17:18:05 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -165.106 uVs 17:18:05 Integrated Voltage Measurement... 17:18:14 Voltage Offset Correction: v_off_ini = -980.600 uV, v_off_fin = -1002.518 uV 17:18:14 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 190.679 uVs 17:18:14 Integrated Voltage Measurement... 17:18:23 Voltage Offset Correction: v_off_ini = -1023.890 uV, v_off_fin = -1003.424 uV 17:18:23 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -175.892 uVs 17:18:23 Integrated Voltage Measurement... 17:18:32 Voltage Offset Correction: v_off_ini = -1000.060 uV, v_off_fin = -992.803 uV 17:18:32 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 208.936 uVs 17:18:34 I2Y Average and RMS of Measurements VT = -189.347 +- 8.386 uVs I2Y = -17.420 +- 0.771 uTm^2 17:18:44 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 17:18:48 Integrated Voltage Measurement... 17:18:57 Voltage Offset Correction: v_off_ini = -976.502 uV, v_off_fin = -972.104 uV 17:18:57 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -125.430 uVs 17:18:57 Integrated Voltage Measurement... 17:19:06 Voltage Offset Correction: v_off_ini = -981.384 uV, v_off_fin = -970.221 uV 17:19:06 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 127.745 uVs 17:19:06 Integrated Voltage Measurement... 17:19:15 Voltage Offset Correction: v_off_ini = -972.789 uV, v_off_fin = -969.647 uV 17:19:15 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -95.015 uVs 17:19:15 Integrated Voltage Measurement... 17:19:24 Voltage Offset Correction: v_off_ini = -986.217 uV, v_off_fin = -985.790 uV 17:19:24 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 93.078 uVs 17:19:24 Integrated Voltage Measurement... 17:19:33 Voltage Offset Correction: v_off_ini = -990.012 uV, v_off_fin = -971.485 uV 17:19:33 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -79.642 uVs 17:19:33 Integrated Voltage Measurement... 17:19:42 Voltage Offset Correction: v_off_ini = -971.923 uV, v_off_fin = -965.632 uV 17:19:42 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 75.384 uVs 17:19:46 I1X Average and RMS of Measurements VT = -99.382 +- 20.387 uVs I1X = 19.877 +- 4.078 uTm 17:19:49 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 17:19:53 Integrated Voltage Measurement... 17:20:02 Voltage Offset Correction: v_off_ini = -952.038 uV, v_off_fin = -980.405 uV 17:20:02 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -189.302 uVs 17:20:02 Integrated Voltage Measurement... 17:20:11 Voltage Offset Correction: v_off_ini = -978.234 uV, v_off_fin = -950.874 uV 17:20:11 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 175.048 uVs 17:20:11 Integrated Voltage Measurement... 17:20:20 Voltage Offset Correction: v_off_ini = -972.934 uV, v_off_fin = -960.775 uV 17:20:20 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -179.282 uVs 17:20:20 Integrated Voltage Measurement... 17:20:29 Voltage Offset Correction: v_off_ini = -964.868 uV, v_off_fin = -961.274 uV 17:20:29 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 176.093 uVs 17:20:29 Integrated Voltage Measurement... 17:20:38 Voltage Offset Correction: v_off_ini = -952.954 uV, v_off_fin = -981.758 uV 17:20:38 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -151.490 uVs 17:20:38 Integrated Voltage Measurement... 17:20:47 Voltage Offset Correction: v_off_ini = -963.012 uV, v_off_fin = -950.387 uV 17:20:47 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 179.234 uVs 17:20:51 I1Y Average and RMS of Measurements VT = -175.075 +- 7.108 uVs I1Y = -35.011 +- 1.421 uTm 17:20:56 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 17:20:59 Integrated Voltage Measurement... 17:21:08 Voltage Offset Correction: v_off_ini = -948.757 uV, v_off_fin = -949.717 uV 17:21:08 Integrated Voltage Measurement dXYus = -1.0002 mm, dXYds = 0.0000 mm dXY = -1.0002 mm VT = -80.170 uVs 17:21:08 Integrated Voltage Measurement... 17:21:17 Voltage Offset Correction: v_off_ini = -955.380 uV, v_off_fin = -950.542 uV 17:21:17 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 78.251 uVs 17:21:17 Integrated Voltage Measurement... 17:21:26 Voltage Offset Correction: v_off_ini = -942.751 uV, v_off_fin = -968.063 uV 17:21:26 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -82.939 uVs 17:21:26 Integrated Voltage Measurement... 17:21:35 Voltage Offset Correction: v_off_ini = -956.004 uV, v_off_fin = -964.837 uV 17:21:35 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 87.496 uVs 17:21:35 Integrated Voltage Measurement... 17:21:44 Voltage Offset Correction: v_off_ini = -955.348 uV, v_off_fin = -961.286 uV 17:21:44 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -89.743 uVs 17:21:44 Integrated Voltage Measurement... 17:21:53 Voltage Offset Correction: v_off_ini = -943.473 uV, v_off_fin = -979.688 uV 17:21:53 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 97.751 uVs 17:21:55 I2X Average and RMS of Measurements VT = -86.058 +- 5.964 uVs I2X = 7.917 +- 0.549 uTm^2 17:22:00 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 17:22:03 Integrated Voltage Measurement... 17:22:12 Voltage Offset Correction: v_off_ini = -967.452 uV, v_off_fin = -963.781 uV 17:22:12 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -231.056 uVs 17:22:12 Integrated Voltage Measurement... 17:22:21 Voltage Offset Correction: v_off_ini = -975.623 uV, v_off_fin = -969.488 uV 17:22:21 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 178.136 uVs 17:22:21 Integrated Voltage Measurement... 17:22:30 Voltage Offset Correction: v_off_ini = -982.655 uV, v_off_fin = -974.048 uV 17:22:30 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -187.721 uVs 17:22:30 Integrated Voltage Measurement... 17:22:39 Voltage Offset Correction: v_off_ini = -967.105 uV, v_off_fin = -985.395 uV 17:22:39 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 215.043 uVs 17:22:39 Integrated Voltage Measurement... 17:22:48 Voltage Offset Correction: v_off_ini = -969.829 uV, v_off_fin = -956.560 uV 17:22:48 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -170.176 uVs 17:22:48 Integrated Voltage Measurement... 17:22:57 Voltage Offset Correction: v_off_ini = -957.234 uV, v_off_fin = -961.017 uV 17:22:57 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 154.660 uVs 17:22:59 I2Y Average and RMS of Measurements VT = -189.465 +- 19.170 uVs I2Y = -17.431 +- 1.764 uTm^2 17:23:09 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 17:23:13 Integrated Voltage Measurement... 17:23:22 Voltage Offset Correction: v_off_ini = -939.019 uV, v_off_fin = -961.604 uV 17:23:22 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -125.558 uVs 17:23:22 Integrated Voltage Measurement... 17:23:31 Voltage Offset Correction: v_off_ini = -931.041 uV, v_off_fin = -970.974 uV 17:23:31 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 83.848 uVs 17:23:31 Integrated Voltage Measurement... 17:23:40 Voltage Offset Correction: v_off_ini = -943.598 uV, v_off_fin = -975.518 uV 17:23:40 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -121.109 uVs 17:23:40 Integrated Voltage Measurement... 17:23:49 Voltage Offset Correction: v_off_ini = -946.078 uV, v_off_fin = -937.076 uV 17:23:49 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 84.429 uVs 17:23:49 Integrated Voltage Measurement... 17:23:58 Voltage Offset Correction: v_off_ini = -983.046 uV, v_off_fin = -956.132 uV 17:23:58 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -80.322 uVs 17:23:58 Integrated Voltage Measurement... 17:24:07 Voltage Offset Correction: v_off_ini = -971.957 uV, v_off_fin = -950.868 uV 17:24:07 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 97.294 uVs 17:24:11 I1X Average and RMS of Measurements VT = -98.760 +- 7.081 uVs I1X = 19.752 +- 1.416 uTm 17:24:14 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 17:24:18 Integrated Voltage Measurement... 17:24:27 Voltage Offset Correction: v_off_ini = -951.489 uV, v_off_fin = -979.761 uV 17:24:27 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -183.532 uVs 17:24:27 Integrated Voltage Measurement... 17:24:36 Voltage Offset Correction: v_off_ini = -962.658 uV, v_off_fin = -921.343 uV 17:24:36 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 176.257 uVs 17:24:36 Integrated Voltage Measurement... 17:24:45 Voltage Offset Correction: v_off_ini = -936.152 uV, v_off_fin = -959.880 uV 17:24:45 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -177.716 uVs 17:24:45 Integrated Voltage Measurement... 17:24:54 Voltage Offset Correction: v_off_ini = -962.079 uV, v_off_fin = -974.114 uV 17:24:54 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 182.646 uVs 17:24:54 Integrated Voltage Measurement... 17:25:03 Voltage Offset Correction: v_off_ini = -985.846 uV, v_off_fin = -959.954 uV 17:25:03 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -160.479 uVs 17:25:03 Integrated Voltage Measurement... 17:25:12 Voltage Offset Correction: v_off_ini = -967.008 uV, v_off_fin = -957.694 uV 17:25:12 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 193.828 uVs 17:25:16 I1Y Average and RMS of Measurements VT = -179.076 +- 1.365 uVs I1Y = -35.812 +- 0.273 uTm 17:25:21 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 17:25:24 Integrated Voltage Measurement... 17:25:33 Voltage Offset Correction: v_off_ini = -979.223 uV, v_off_fin = -983.738 uV 17:25:33 Integrated Voltage Measurement dXYus = -1.0002 mm, dXYds = 0.0000 mm dXY = -1.0002 mm VT = -79.474 uVs 17:25:33 Integrated Voltage Measurement... 17:25:42 Voltage Offset Correction: v_off_ini = -942.828 uV, v_off_fin = -962.784 uV 17:25:42 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 87.736 uVs 17:25:42 Integrated Voltage Measurement... 17:25:51 Voltage Offset Correction: v_off_ini = -985.068 uV, v_off_fin = -982.307 uV 17:25:51 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -82.625 uVs 17:25:51 Integrated Voltage Measurement... 17:26:00 Voltage Offset Correction: v_off_ini = -973.720 uV, v_off_fin = -969.270 uV 17:26:00 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 109.570 uVs 17:26:00 Integrated Voltage Measurement... 17:26:09 Voltage Offset Correction: v_off_ini = -971.825 uV, v_off_fin = -962.647 uV 17:26:09 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -47.920 uVs 17:26:09 Integrated Voltage Measurement... 17:26:18 Voltage Offset Correction: v_off_ini = -991.018 uV, v_off_fin = -1002.542 uV 17:26:18 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 70.178 uVs 17:26:20 I2X Average and RMS of Measurements VT = -79.584 +- 15.390 uVs I2X = 7.322 +- 1.416 uTm^2 17:26:25 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 17:26:28 Integrated Voltage Measurement... 17:26:37 Voltage Offset Correction: v_off_ini = -960.464 uV, v_off_fin = -979.587 uV 17:26:37 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -207.421 uVs 17:26:37 Integrated Voltage Measurement... 17:26:46 Voltage Offset Correction: v_off_ini = -1015.185 uV, v_off_fin = -977.971 uV 17:26:46 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 189.759 uVs 17:26:46 Integrated Voltage Measurement... 17:26:55 Voltage Offset Correction: v_off_ini = -1001.152 uV, v_off_fin = -972.626 uV 17:26:55 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -166.017 uVs 17:26:55 Integrated Voltage Measurement... 17:27:04 Voltage Offset Correction: v_off_ini = -994.352 uV, v_off_fin = -1011.415 uV 17:27:04 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 200.712 uVs 17:27:04 Integrated Voltage Measurement... 17:27:13 Voltage Offset Correction: v_off_ini = -1003.881 uV, v_off_fin = -1009.649 uV 17:27:13 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -184.976 uVs 17:27:13 Integrated Voltage Measurement... 17:27:22 Voltage Offset Correction: v_off_ini = -1004.490 uV, v_off_fin = -999.953 uV 17:27:22 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 174.599 uVs 17:27:24 I2Y Average and RMS of Measurements VT = -187.247 +- 8.152 uVs I2Y = -17.227 +- 0.750 uTm^2 17:27:34 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 17:27:38 Integrated Voltage Measurement... 17:27:47 Voltage Offset Correction: v_off_ini = -997.985 uV, v_off_fin = -1000.876 uV 17:27:47 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.4998 mm dXY = -0.4999 mm VT = -100.433 uVs 17:27:47 Integrated Voltage Measurement... 17:27:56 Voltage Offset Correction: v_off_ini = -1022.555 uV, v_off_fin = -995.853 uV 17:27:56 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 96.331 uVs 17:27:56 Integrated Voltage Measurement... 17:28:05 Voltage Offset Correction: v_off_ini = -997.780 uV, v_off_fin = -1015.372 uV 17:28:05 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -89.299 uVs 17:28:05 Integrated Voltage Measurement... 17:28:14 Voltage Offset Correction: v_off_ini = -1022.521 uV, v_off_fin = -1027.058 uV 17:28:14 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 66.109 uVs 17:28:14 Integrated Voltage Measurement... 17:28:23 Voltage Offset Correction: v_off_ini = -1024.929 uV, v_off_fin = -1021.769 uV 17:28:23 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -63.362 uVs 17:28:23 Integrated Voltage Measurement... 17:28:32 Voltage Offset Correction: v_off_ini = -1050.325 uV, v_off_fin = -1000.378 uV 17:28:32 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 77.098 uVs 17:28:36 I1X Average and RMS of Measurements VT = -82.105 +- 11.907 uVs I1X = 16.422 +- 2.383 uTm 17:28:39 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 17:28:43 Integrated Voltage Measurement... 17:28:52 Voltage Offset Correction: v_off_ini = -1020.861 uV, v_off_fin = -1031.223 uV 17:28:52 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -184.842 uVs 17:28:52 Integrated Voltage Measurement... 17:29:01 Voltage Offset Correction: v_off_ini = -1025.525 uV, v_off_fin = -1001.076 uV 17:29:01 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 168.614 uVs 17:29:01 Integrated Voltage Measurement... 17:29:10 Voltage Offset Correction: v_off_ini = -985.874 uV, v_off_fin = -1017.887 uV 17:29:10 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -175.198 uVs 17:29:10 Integrated Voltage Measurement... 17:29:19 Voltage Offset Correction: v_off_ini = -1037.289 uV, v_off_fin = -1004.171 uV 17:29:19 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 182.796 uVs 17:29:19 Integrated Voltage Measurement... 17:29:28 Voltage Offset Correction: v_off_ini = -1036.507 uV, v_off_fin = -1037.987 uV 17:29:28 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -184.410 uVs 17:29:28 Integrated Voltage Measurement... 17:29:37 Voltage Offset Correction: v_off_ini = -1027.395 uV, v_off_fin = -1026.011 uV 17:29:37 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 188.038 uVs 17:29:41 I1Y Average and RMS of Measurements VT = -180.650 +- 4.049 uVs I1Y = -36.126 +- 0.810 uTm 17:29:46 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 17:29:49 Integrated Voltage Measurement... 17:29:58 Voltage Offset Correction: v_off_ini = -1017.699 uV, v_off_fin = -1013.128 uV 17:29:58 Integrated Voltage Measurement dXYus = -1.0001 mm, dXYds = 0.0000 mm dXY = -1.0001 mm VT = -98.372 uVs 17:29:58 Integrated Voltage Measurement... 17:30:07 Voltage Offset Correction: v_off_ini = -1019.291 uV, v_off_fin = -1025.675 uV 17:30:07 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 72.094 uVs 17:30:07 Integrated Voltage Measurement... 17:30:16 Voltage Offset Correction: v_off_ini = -1052.460 uV, v_off_fin = -1038.775 uV 17:30:16 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -85.964 uVs 17:30:16 Integrated Voltage Measurement... 17:30:25 Voltage Offset Correction: v_off_ini = -1037.835 uV, v_off_fin = -1030.458 uV 17:30:25 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 73.140 uVs 17:30:25 Integrated Voltage Measurement... 17:30:34 Voltage Offset Correction: v_off_ini = -993.965 uV, v_off_fin = -1029.776 uV 17:30:34 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -73.116 uVs 17:30:34 Integrated Voltage Measurement... 17:30:43 Voltage Offset Correction: v_off_ini = -989.800 uV, v_off_fin = -1009.785 uV 17:30:43 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 67.227 uVs 17:30:45 I2X Average and RMS of Measurements VT = -78.319 +- 6.210 uVs I2X = 7.205 +- 0.571 uTm^2 17:30:50 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 17:30:53 Integrated Voltage Measurement... 17:31:02 Voltage Offset Correction: v_off_ini = -1008.350 uV, v_off_fin = -1026.556 uV 17:31:02 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -163.786 uVs 17:31:02 Integrated Voltage Measurement... 17:31:11 Voltage Offset Correction: v_off_ini = -1017.391 uV, v_off_fin = -1022.327 uV 17:31:11 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 184.013 uVs 17:31:11 Integrated Voltage Measurement... 17:31:20 Voltage Offset Correction: v_off_ini = -1027.976 uV, v_off_fin = -1017.066 uV 17:31:20 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -154.957 uVs 17:31:20 Integrated Voltage Measurement... 17:31:29 Voltage Offset Correction: v_off_ini = -1004.620 uV, v_off_fin = -993.543 uV 17:31:29 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 156.877 uVs 17:31:29 Integrated Voltage Measurement... 17:31:38 Voltage Offset Correction: v_off_ini = -1018.801 uV, v_off_fin = -1018.568 uV 17:31:38 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -168.881 uVs 17:31:38 Integrated Voltage Measurement... 17:31:47 Voltage Offset Correction: v_off_ini = -1011.848 uV, v_off_fin = -1009.838 uV 17:31:47 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 160.163 uVs 17:31:49 I2Y Average and RMS of Measurements VT = -164.779 +- 7.344 uVs I2Y = -15.160 +- 0.676 uTm^2 17:32:01 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 17:32:05 Integrated Voltage Measurement... 17:32:14 Voltage Offset Correction: v_off_ini = -999.318 uV, v_off_fin = -993.069 uV 17:32:14 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -75.296 uVs 17:32:14 Integrated Voltage Measurement... 17:32:23 Voltage Offset Correction: v_off_ini = -1010.678 uV, v_off_fin = -1027.991 uV 17:32:23 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 89.183 uVs 17:32:23 Integrated Voltage Measurement... 17:32:32 Voltage Offset Correction: v_off_ini = -999.182 uV, v_off_fin = -1004.862 uV 17:32:32 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -74.012 uVs 17:32:32 Integrated Voltage Measurement... 17:32:41 Voltage Offset Correction: v_off_ini = -988.201 uV, v_off_fin = -995.600 uV 17:32:41 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 96.172 uVs 17:32:41 Integrated Voltage Measurement... 17:32:50 Voltage Offset Correction: v_off_ini = -968.129 uV, v_off_fin = -985.865 uV 17:32:50 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -81.238 uVs 17:32:50 Integrated Voltage Measurement... 17:32:59 Voltage Offset Correction: v_off_ini = -987.035 uV, v_off_fin = -993.836 uV 17:32:59 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 84.548 uVs 17:33:03 I1X Average and RMS of Measurements VT = -83.408 +- 1.220 uVs I1X = 16.682 +- 0.244 uTm 17:33:06 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 17:33:10 Integrated Voltage Measurement... 17:33:19 Voltage Offset Correction: v_off_ini = -1020.429 uV, v_off_fin = -994.484 uV 17:33:19 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -161.020 uVs 17:33:19 Integrated Voltage Measurement... 17:33:28 Voltage Offset Correction: v_off_ini = -981.063 uV, v_off_fin = -980.186 uV 17:33:28 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 148.415 uVs 17:33:28 Integrated Voltage Measurement... 17:33:37 Voltage Offset Correction: v_off_ini = -973.986 uV, v_off_fin = -978.050 uV 17:33:37 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -147.140 uVs 17:33:37 Integrated Voltage Measurement... 17:33:46 Voltage Offset Correction: v_off_ini = -1013.576 uV, v_off_fin = -982.397 uV 17:33:46 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 157.220 uVs 17:33:46 Integrated Voltage Measurement... 17:33:55 Voltage Offset Correction: v_off_ini = -961.581 uV, v_off_fin = -1001.559 uV 17:33:55 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -161.660 uVs 17:33:55 Integrated Voltage Measurement... 17:34:04 Voltage Offset Correction: v_off_ini = -996.284 uV, v_off_fin = -986.560 uV 17:34:04 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 167.284 uVs 17:34:08 I1Y Average and RMS of Measurements VT = -157.123 +- 5.298 uVs I1Y = -31.423 +- 1.061 uTm 17:34:13 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 17:34:16 Integrated Voltage Measurement... 17:34:25 Voltage Offset Correction: v_off_ini = -988.614 uV, v_off_fin = -982.809 uV 17:34:25 Integrated Voltage Measurement dXYus = -1.0002 mm, dXYds = 0.0000 mm dXY = -1.0002 mm VT = -87.126 uVs 17:34:25 Integrated Voltage Measurement... 17:34:34 Voltage Offset Correction: v_off_ini = -994.187 uV, v_off_fin = -989.392 uV 17:34:34 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 51.362 uVs 17:34:34 Integrated Voltage Measurement... 17:34:43 Voltage Offset Correction: v_off_ini = -1014.473 uV, v_off_fin = -1006.638 uV 17:34:43 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -57.322 uVs 17:34:43 Integrated Voltage Measurement... 17:34:52 Voltage Offset Correction: v_off_ini = -993.320 uV, v_off_fin = -992.380 uV 17:34:52 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 65.639 uVs 17:34:52 Integrated Voltage Measurement... 17:35:01 Voltage Offset Correction: v_off_ini = -975.499 uV, v_off_fin = -977.047 uV 17:35:01 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -107.881 uVs 17:35:01 Integrated Voltage Measurement... 17:35:10 Voltage Offset Correction: v_off_ini = -1013.367 uV, v_off_fin = -1005.493 uV 17:35:10 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 62.485 uVs 17:35:12 I2X Average and RMS of Measurements VT = -71.969 +- 9.867 uVs I2X = 6.621 +- 0.908 uTm^2 17:35:17 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 17:35:20 Integrated Voltage Measurement... 17:35:29 Voltage Offset Correction: v_off_ini = -995.634 uV, v_off_fin = -990.506 uV 17:35:29 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -173.131 uVs 17:35:29 Integrated Voltage Measurement... 17:35:38 Voltage Offset Correction: v_off_ini = -1008.229 uV, v_off_fin = -993.526 uV 17:35:38 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 156.175 uVs 17:35:38 Integrated Voltage Measurement... 17:35:47 Voltage Offset Correction: v_off_ini = -1028.737 uV, v_off_fin = -994.350 uV 17:35:47 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -157.428 uVs 17:35:47 Integrated Voltage Measurement... 17:35:56 Voltage Offset Correction: v_off_ini = -1031.473 uV, v_off_fin = -1011.036 uV 17:35:56 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 178.709 uVs 17:35:56 Integrated Voltage Measurement... 17:36:05 Voltage Offset Correction: v_off_ini = -1018.753 uV, v_off_fin = -992.938 uV 17:36:05 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -141.622 uVs 17:36:05 Integrated Voltage Measurement... 17:36:14 Voltage Offset Correction: v_off_ini = -989.243 uV, v_off_fin = -997.179 uV 17:36:14 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 136.321 uVs 17:36:16 I2Y Average and RMS of Measurements VT = -157.231 +- 12.986 uVs I2Y = -14.465 +- 1.195 uTm^2 17:36:27 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 17:36:31 Integrated Voltage Measurement... 17:36:40 Voltage Offset Correction: v_off_ini = -984.008 uV, v_off_fin = -989.887 uV 17:36:40 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.4997 mm dXY = -0.4998 mm VT = -53.754 uVs 17:36:40 Integrated Voltage Measurement... 17:36:49 Voltage Offset Correction: v_off_ini = -993.305 uV, v_off_fin = -985.329 uV 17:36:49 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = 72.969 uVs 17:36:49 Integrated Voltage Measurement... 17:36:58 Voltage Offset Correction: v_off_ini = -988.339 uV, v_off_fin = -987.913 uV 17:36:58 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -65.462 uVs 17:36:58 Integrated Voltage Measurement... 17:37:07 Voltage Offset Correction: v_off_ini = -994.148 uV, v_off_fin = -970.963 uV 17:37:07 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 52.016 uVs 17:37:07 Integrated Voltage Measurement... 17:37:16 Voltage Offset Correction: v_off_ini = -974.926 uV, v_off_fin = -981.003 uV 17:37:16 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -70.021 uVs 17:37:16 Integrated Voltage Measurement... 17:37:25 Voltage Offset Correction: v_off_ini = -970.122 uV, v_off_fin = -989.440 uV 17:37:25 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 80.313 uVs 17:37:29 I1X Average and RMS of Measurements VT = -65.756 +- 6.917 uVs I1X = 13.152 +- 1.383 uTm 17:37:32 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 17:37:36 Integrated Voltage Measurement... 17:37:45 Voltage Offset Correction: v_off_ini = -984.418 uV, v_off_fin = -975.010 uV 17:37:45 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -160.009 uVs 17:37:45 Integrated Voltage Measurement... 17:37:54 Voltage Offset Correction: v_off_ini = -962.664 uV, v_off_fin = -952.909 uV 17:37:54 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 147.579 uVs 17:37:54 Integrated Voltage Measurement... 17:38:03 Voltage Offset Correction: v_off_ini = -946.809 uV, v_off_fin = -967.132 uV 17:38:03 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -135.570 uVs 17:38:03 Integrated Voltage Measurement... 17:38:12 Voltage Offset Correction: v_off_ini = -950.460 uV, v_off_fin = -933.081 uV 17:38:12 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = 150.170 uVs 17:38:12 Integrated Voltage Measurement... 17:38:21 Voltage Offset Correction: v_off_ini = -942.182 uV, v_off_fin = -961.745 uV 17:38:21 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -155.482 uVs 17:38:21 Integrated Voltage Measurement... 17:38:30 Voltage Offset Correction: v_off_ini = -954.717 uV, v_off_fin = -950.814 uV 17:38:30 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 176.730 uVs 17:38:34 I1Y Average and RMS of Measurements VT = -154.257 +- 9.492 uVs I1Y = -30.850 +- 1.897 uTm 17:38:39 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 17:38:42 Integrated Voltage Measurement... 17:38:51 Voltage Offset Correction: v_off_ini = -971.482 uV, v_off_fin = -965.705 uV 17:38:51 Integrated Voltage Measurement dXYus = -1.0001 mm, dXYds = 0.0000 mm dXY = -1.0001 mm VT = -69.943 uVs 17:38:51 Integrated Voltage Measurement... 17:39:00 Voltage Offset Correction: v_off_ini = -963.374 uV, v_off_fin = -966.719 uV 17:39:00 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 87.700 uVs 17:39:00 Integrated Voltage Measurement... 17:39:09 Voltage Offset Correction: v_off_ini = -958.203 uV, v_off_fin = -952.113 uV 17:39:09 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -88.945 uVs 17:39:09 Integrated Voltage Measurement... 17:39:18 Voltage Offset Correction: v_off_ini = -977.101 uV, v_off_fin = -971.614 uV 17:39:18 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 72.936 uVs 17:39:18 Integrated Voltage Measurement... 17:39:27 Voltage Offset Correction: v_off_ini = -955.594 uV, v_off_fin = -978.993 uV 17:39:27 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -79.787 uVs 17:39:27 Integrated Voltage Measurement... 17:39:36 Voltage Offset Correction: v_off_ini = -986.690 uV, v_off_fin = -958.385 uV 17:39:36 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 50.117 uVs 17:39:38 I2X Average and RMS of Measurements VT = -74.905 +- 7.091 uVs I2X = 6.891 +- 0.652 uTm^2 17:39:43 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 17:39:46 Integrated Voltage Measurement... 17:39:55 Voltage Offset Correction: v_off_ini = -954.242 uV, v_off_fin = -950.594 uV 17:39:55 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -135.083 uVs 17:39:55 Integrated Voltage Measurement... 17:40:04 Voltage Offset Correction: v_off_ini = -951.455 uV, v_off_fin = -960.796 uV 17:40:04 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 173.862 uVs 17:40:04 Integrated Voltage Measurement... 17:40:13 Voltage Offset Correction: v_off_ini = -952.431 uV, v_off_fin = -967.105 uV 17:40:13 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -161.603 uVs 17:40:13 Integrated Voltage Measurement... 17:40:22 Voltage Offset Correction: v_off_ini = -984.401 uV, v_off_fin = -981.246 uV 17:40:22 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 152.613 uVs 17:40:22 Integrated Voltage Measurement... 17:40:31 Voltage Offset Correction: v_off_ini = -958.929 uV, v_off_fin = -965.479 uV 17:40:31 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -163.487 uVs 17:40:31 Integrated Voltage Measurement... 17:40:40 Voltage Offset Correction: v_off_ini = -938.487 uV, v_off_fin = -950.200 uV 17:40:40 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 147.446 uVs 17:40:42 I2Y Average and RMS of Measurements VT = -155.682 +- 1.087 uVs I2Y = -14.323 +- 0.100 uTm^2 17:40:55 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 17:40:59 Integrated Voltage Measurement... 17:41:08 Voltage Offset Correction: v_off_ini = -941.823 uV, v_off_fin = -942.615 uV 17:41:08 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -46.619 uVs 17:41:08 Integrated Voltage Measurement... 17:41:17 Voltage Offset Correction: v_off_ini = -952.020 uV, v_off_fin = -948.941 uV 17:41:17 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 79.831 uVs 17:41:17 Integrated Voltage Measurement... 17:41:26 Voltage Offset Correction: v_off_ini = -944.777 uV, v_off_fin = -939.105 uV 17:41:26 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -61.032 uVs 17:41:26 Integrated Voltage Measurement... 17:41:35 Voltage Offset Correction: v_off_ini = -955.051 uV, v_off_fin = -946.708 uV 17:41:35 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 31.349 uVs 17:41:35 Integrated Voltage Measurement... 17:41:44 Voltage Offset Correction: v_off_ini = -954.478 uV, v_off_fin = -962.820 uV 17:41:44 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -46.351 uVs 17:41:44 Integrated Voltage Measurement... 17:41:53 Voltage Offset Correction: v_off_ini = -947.190 uV, v_off_fin = -950.336 uV 17:41:53 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 47.808 uVs 17:41:58 I1X Average and RMS of Measurements VT = -52.165 +- 7.829 uVs I1X = 10.433 +- 1.566 uTm 17:42:00 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 17:42:04 Integrated Voltage Measurement... 17:42:13 Voltage Offset Correction: v_off_ini = -941.455 uV, v_off_fin = -946.926 uV 17:42:13 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -142.065 uVs 17:42:13 Integrated Voltage Measurement... 17:42:22 Voltage Offset Correction: v_off_ini = -923.705 uV, v_off_fin = -947.022 uV 17:42:22 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 159.299 uVs 17:42:22 Integrated Voltage Measurement... 17:42:31 Voltage Offset Correction: v_off_ini = -969.184 uV, v_off_fin = -960.790 uV 17:42:31 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -151.753 uVs 17:42:31 Integrated Voltage Measurement... 17:42:40 Voltage Offset Correction: v_off_ini = -935.192 uV, v_off_fin = -940.945 uV 17:42:40 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = 151.705 uVs 17:42:40 Integrated Voltage Measurement... 17:42:49 Voltage Offset Correction: v_off_ini = -972.915 uV, v_off_fin = -948.069 uV 17:42:49 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -144.146 uVs 17:42:49 Integrated Voltage Measurement... 17:42:58 Voltage Offset Correction: v_off_ini = -959.259 uV, v_off_fin = -945.294 uV 17:42:58 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 164.589 uVs 17:43:02 I1Y Average and RMS of Measurements VT = -152.260 +- 1.551 uVs I1Y = -30.450 +- 0.309 uTm 17:43:07 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 17:43:10 Integrated Voltage Measurement... 17:43:19 Voltage Offset Correction: v_off_ini = -930.515 uV, v_off_fin = -934.809 uV 17:43:19 Integrated Voltage Measurement dXYus = -1.0002 mm, dXYds = 0.0000 mm dXY = -1.0002 mm VT = -75.016 uVs 17:43:19 Integrated Voltage Measurement... 17:43:28 Voltage Offset Correction: v_off_ini = -955.048 uV, v_off_fin = -942.127 uV 17:43:28 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 58.249 uVs 17:43:28 Integrated Voltage Measurement... 17:43:37 Voltage Offset Correction: v_off_ini = -932.597 uV, v_off_fin = -940.410 uV 17:43:37 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -48.479 uVs 17:43:37 Integrated Voltage Measurement... 17:43:46 Voltage Offset Correction: v_off_ini = -946.709 uV, v_off_fin = -957.945 uV 17:43:46 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 67.822 uVs 17:43:46 Integrated Voltage Measurement... 17:43:55 Voltage Offset Correction: v_off_ini = -924.205 uV, v_off_fin = -951.605 uV 17:43:55 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -80.471 uVs 17:43:55 Integrated Voltage Measurement... 17:44:04 Voltage Offset Correction: v_off_ini = -944.924 uV, v_off_fin = -937.898 uV 17:44:04 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 50.026 uVs 17:44:06 I2X Average and RMS of Measurements VT = -63.344 +- 3.716 uVs I2X = 5.827 +- 0.342 uTm^2 17:44:11 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 17:44:14 Integrated Voltage Measurement... 17:44:23 Voltage Offset Correction: v_off_ini = -924.801 uV, v_off_fin = -936.284 uV 17:44:23 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -140.719 uVs 17:44:23 Integrated Voltage Measurement... 17:44:32 Voltage Offset Correction: v_off_ini = -933.576 uV, v_off_fin = -922.239 uV 17:44:32 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 142.860 uVs 17:44:32 Integrated Voltage Measurement... 17:44:41 Voltage Offset Correction: v_off_ini = -915.853 uV, v_off_fin = -933.210 uV 17:44:41 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -137.989 uVs 17:44:41 Integrated Voltage Measurement... 17:44:50 Voltage Offset Correction: v_off_ini = -932.912 uV, v_off_fin = -933.155 uV 17:44:50 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 159.935 uVs 17:44:50 Integrated Voltage Measurement... 17:44:59 Voltage Offset Correction: v_off_ini = -940.293 uV, v_off_fin = -939.180 uV 17:44:59 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -154.035 uVs 17:44:59 Integrated Voltage Measurement... 17:45:08 Voltage Offset Correction: v_off_ini = -940.884 uV, v_off_fin = -938.044 uV 17:45:08 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 129.382 uVs 17:45:10 I2Y Average and RMS of Measurements VT = -144.153 +- 3.401 uVs I2Y = -13.262 +- 0.313 uTm^2 17:45:23 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 17:45:28 Integrated Voltage Measurement... 17:45:37 Voltage Offset Correction: v_off_ini = -921.408 uV, v_off_fin = -919.127 uV 17:45:37 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.4998 mm dXY = -0.4999 mm VT = -35.228 uVs 17:45:37 Integrated Voltage Measurement... 17:45:46 Voltage Offset Correction: v_off_ini = -964.946 uV, v_off_fin = -931.600 uV 17:45:46 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 45.134 uVs 17:45:46 Integrated Voltage Measurement... 17:45:55 Voltage Offset Correction: v_off_ini = -932.266 uV, v_off_fin = -924.570 uV 17:45:55 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -46.749 uVs 17:45:55 Integrated Voltage Measurement... 17:46:04 Voltage Offset Correction: v_off_ini = -926.160 uV, v_off_fin = -954.962 uV 17:46:04 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 11.613 uVs 17:46:04 Integrated Voltage Measurement... 17:46:13 Voltage Offset Correction: v_off_ini = -932.515 uV, v_off_fin = -939.777 uV 17:46:13 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -38.709 uVs 17:46:13 Integrated Voltage Measurement... 17:46:22 Voltage Offset Correction: v_off_ini = -943.337 uV, v_off_fin = -945.036 uV 17:46:22 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 56.898 uVs 17:46:26 I1X Average and RMS of Measurements VT = -39.055 +- 7.644 uVs I1X = 7.811 +- 1.529 uTm 17:46:28 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 17:46:33 Integrated Voltage Measurement... 17:46:42 Voltage Offset Correction: v_off_ini = -935.686 uV, v_off_fin = -941.304 uV 17:46:42 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -123.632 uVs 17:46:42 Integrated Voltage Measurement... 17:46:50 Voltage Offset Correction: v_off_ini = -930.820 uV, v_off_fin = -939.693 uV 17:46:51 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = 168.833 uVs 17:46:51 Integrated Voltage Measurement... 17:46:59 Voltage Offset Correction: v_off_ini = -937.275 uV, v_off_fin = -956.036 uV 17:46:59 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5002 mm dXY = 0.5001 mm VT = -149.421 uVs 17:46:59 Integrated Voltage Measurement... 17:47:08 Voltage Offset Correction: v_off_ini = -952.949 uV, v_off_fin = -949.764 uV 17:47:08 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5002 mm dXY = -0.5001 mm VT = 142.866 uVs 17:47:08 Integrated Voltage Measurement... 17:47:17 Voltage Offset Correction: v_off_ini = -930.537 uV, v_off_fin = -962.071 uV 17:47:17 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.5002 mm dXY = 0.5000 mm VT = -133.898 uVs 17:47:17 Integrated Voltage Measurement... 17:47:26 Voltage Offset Correction: v_off_ini = -947.516 uV, v_off_fin = -954.305 uV 17:47:26 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5002 mm dXY = -0.5001 mm VT = 120.162 uVs 17:47:31 I1Y Average and RMS of Measurements VT = -139.802 +- 9.031 uVs I1Y = -27.957 +- 1.807 uTm 17:47:36 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 17:47:38 Integrated Voltage Measurement... 17:47:47 Voltage Offset Correction: v_off_ini = -934.557 uV, v_off_fin = -942.465 uV 17:47:47 Integrated Voltage Measurement dXYus = -1.0002 mm, dXYds = 0.0000 mm dXY = -1.0002 mm VT = -61.402 uVs 17:47:47 Integrated Voltage Measurement... 17:47:56 Voltage Offset Correction: v_off_ini = -932.277 uV, v_off_fin = -943.635 uV 17:47:56 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 81.727 uVs 17:47:56 Integrated Voltage Measurement... 17:48:05 Voltage Offset Correction: v_off_ini = -948.058 uV, v_off_fin = -962.342 uV 17:48:05 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -57.253 uVs 17:48:05 Integrated Voltage Measurement... 17:48:14 Voltage Offset Correction: v_off_ini = -936.691 uV, v_off_fin = -946.408 uV 17:48:14 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 21.624 uVs 17:48:14 Integrated Voltage Measurement... 17:48:23 Voltage Offset Correction: v_off_ini = -949.372 uV, v_off_fin = -964.929 uV 17:48:23 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -4.591 uVs 17:48:23 Integrated Voltage Measurement... 17:48:32 Voltage Offset Correction: v_off_ini = -952.544 uV, v_off_fin = -965.000 uV 17:48:32 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 70.161 uVs 17:48:35 I2X Average and RMS of Measurements VT = -49.460 +- 15.653 uVs I2X = 4.550 +- 1.440 uTm^2 17:48:40 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 17:48:42 Integrated Voltage Measurement... 17:48:51 Voltage Offset Correction: v_off_ini = -939.520 uV, v_off_fin = -940.789 uV 17:48:51 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -129.667 uVs 17:48:51 Integrated Voltage Measurement... 17:49:00 Voltage Offset Correction: v_off_ini = -924.957 uV, v_off_fin = -940.195 uV 17:49:00 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 140.966 uVs 17:49:00 Integrated Voltage Measurement... 17:49:09 Voltage Offset Correction: v_off_ini = -906.820 uV, v_off_fin = -936.938 uV 17:49:09 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0001 mm dXY = 1.0000 mm VT = -120.784 uVs 17:49:09 Integrated Voltage Measurement... 17:49:18 Voltage Offset Correction: v_off_ini = -942.853 uV, v_off_fin = -919.831 uV 17:49:18 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 130.645 uVs 17:49:18 Integrated Voltage Measurement... 17:49:27 Voltage Offset Correction: v_off_ini = -924.929 uV, v_off_fin = -930.247 uV 17:49:27 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -160.453 uVs 17:49:27 Integrated Voltage Measurement... 17:49:36 Voltage Offset Correction: v_off_ini = -940.994 uV, v_off_fin = -939.172 uV 17:49:36 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 126.590 uVs 17:49:39 I2Y Average and RMS of Measurements VT = -134.851 +- 7.277 uVs I2Y = -12.406 +- 0.669 uTm^2 17:49:52 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 17:49:56 Integrated Voltage Measurement... 17:50:05 Voltage Offset Correction: v_off_ini = -949.126 uV, v_off_fin = -945.411 uV 17:50:05 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4998 mm dXY = -0.4999 mm VT = -46.148 uVs 17:50:05 Integrated Voltage Measurement... 17:50:14 Voltage Offset Correction: v_off_ini = -950.286 uV, v_off_fin = -946.917 uV 17:50:14 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 38.239 uVs 17:50:14 Integrated Voltage Measurement... 17:50:23 Voltage Offset Correction: v_off_ini = -943.291 uV, v_off_fin = -957.455 uV 17:50:23 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -44.405 uVs 17:50:23 Integrated Voltage Measurement... 17:50:32 Voltage Offset Correction: v_off_ini = -964.237 uV, v_off_fin = -944.924 uV 17:50:32 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 54.257 uVs 17:50:32 Integrated Voltage Measurement... 17:50:41 Voltage Offset Correction: v_off_ini = -951.364 uV, v_off_fin = -928.343 uV 17:50:41 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -61.494 uVs 17:50:41 Integrated Voltage Measurement... 17:50:50 Voltage Offset Correction: v_off_ini = -938.152 uV, v_off_fin = -937.344 uV 17:50:50 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 39.888 uVs 17:50:54 I1X Average and RMS of Measurements VT = -47.405 +- 3.727 uVs I1X = 9.481 +- 0.745 uTm 17:50:57 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 17:51:01 Integrated Voltage Measurement... 17:51:10 Voltage Offset Correction: v_off_ini = -933.276 uV, v_off_fin = -931.926 uV 17:51:10 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -150.504 uVs 17:51:10 Integrated Voltage Measurement... 17:51:19 Voltage Offset Correction: v_off_ini = -940.472 uV, v_off_fin = -936.447 uV 17:51:19 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 140.553 uVs 17:51:19 Integrated Voltage Measurement... 17:51:28 Voltage Offset Correction: v_off_ini = -938.562 uV, v_off_fin = -950.377 uV 17:51:28 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -171.034 uVs 17:51:28 Integrated Voltage Measurement... 17:51:37 Voltage Offset Correction: v_off_ini = -943.948 uV, v_off_fin = -932.567 uV 17:51:37 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 143.086 uVs 17:51:37 Integrated Voltage Measurement... 17:51:46 Voltage Offset Correction: v_off_ini = -935.782 uV, v_off_fin = -933.941 uV 17:51:46 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -156.345 uVs 17:51:46 Integrated Voltage Measurement... 17:51:55 Voltage Offset Correction: v_off_ini = -922.328 uV, v_off_fin = -916.772 uV 17:51:55 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 152.460 uVs 17:51:59 I1Y Average and RMS of Measurements VT = -152.330 +- 4.931 uVs I1Y = -30.464 +- 0.986 uTm 17:52:04 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 17:52:06 Integrated Voltage Measurement... 17:52:15 Voltage Offset Correction: v_off_ini = -928.077 uV, v_off_fin = -936.479 uV 17:52:15 Integrated Voltage Measurement dXYus = -1.0002 mm, dXYds = 0.0000 mm dXY = -1.0002 mm VT = -68.103 uVs 17:52:16 Integrated Voltage Measurement... 17:52:24 Voltage Offset Correction: v_off_ini = -925.689 uV, v_off_fin = -948.285 uV 17:52:25 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 47.247 uVs 17:52:25 Integrated Voltage Measurement... 17:52:33 Voltage Offset Correction: v_off_ini = -937.222 uV, v_off_fin = -960.219 uV 17:52:34 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -13.695 uVs 17:52:34 Integrated Voltage Measurement... 17:52:42 Voltage Offset Correction: v_off_ini = -939.893 uV, v_off_fin = -934.268 uV 17:52:43 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 42.340 uVs 17:52:43 Integrated Voltage Measurement... 17:52:51 Voltage Offset Correction: v_off_ini = -930.251 uV, v_off_fin = -941.619 uV 17:52:52 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -26.889 uVs 17:52:52 Integrated Voltage Measurement... 17:53:00 Voltage Offset Correction: v_off_ini = -922.599 uV, v_off_fin = -922.075 uV 17:53:01 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 57.190 uVs 17:53:03 I2X Average and RMS of Measurements VT = -42.577 +- 12.114 uVs I2X = 3.917 +- 1.114 uTm^2 17:53:08 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 17:53:11 Integrated Voltage Measurement... 17:53:20 Voltage Offset Correction: v_off_ini = -933.436 uV, v_off_fin = -916.333 uV 17:53:20 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -136.453 uVs 17:53:20 Integrated Voltage Measurement... 17:53:29 Voltage Offset Correction: v_off_ini = -924.912 uV, v_off_fin = -924.494 uV 17:53:29 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 140.411 uVs 17:53:29 Integrated Voltage Measurement... 17:53:38 Voltage Offset Correction: v_off_ini = -935.966 uV, v_off_fin = -903.711 uV 17:53:38 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -142.522 uVs 17:53:38 Integrated Voltage Measurement... 17:53:47 Voltage Offset Correction: v_off_ini = -924.795 uV, v_off_fin = -960.683 uV 17:53:47 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 165.789 uVs 17:53:47 Integrated Voltage Measurement... 17:53:56 Voltage Offset Correction: v_off_ini = -927.678 uV, v_off_fin = -922.434 uV 17:53:56 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -158.519 uVs 17:53:56 Integrated Voltage Measurement... 17:54:05 Voltage Offset Correction: v_off_ini = -907.775 uV, v_off_fin = -931.973 uV 17:54:05 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 125.520 uVs 17:54:07 I2Y Average and RMS of Measurements VT = -144.869 +- 6.728 uVs I2Y = -13.328 +- 0.619 uTm^2 17:54:20 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 17:54:24 Integrated Voltage Measurement... 17:54:33 Voltage Offset Correction: v_off_ini = -901.287 uV, v_off_fin = -914.522 uV 17:54:33 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -26.411 uVs 17:54:33 Integrated Voltage Measurement... 17:54:42 Voltage Offset Correction: v_off_ini = -919.453 uV, v_off_fin = -920.248 uV 17:54:42 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 54.895 uVs 17:54:42 Integrated Voltage Measurement... 17:54:51 Voltage Offset Correction: v_off_ini = -914.413 uV, v_off_fin = -921.896 uV 17:54:51 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -19.351 uVs 17:54:51 Integrated Voltage Measurement... 17:55:00 Voltage Offset Correction: v_off_ini = -913.989 uV, v_off_fin = -922.396 uV 17:55:00 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 23.682 uVs 17:55:00 Integrated Voltage Measurement... 17:55:09 Voltage Offset Correction: v_off_ini = -910.405 uV, v_off_fin = -930.279 uV 17:55:09 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -16.840 uVs 17:55:09 Integrated Voltage Measurement... 17:55:18 Voltage Offset Correction: v_off_ini = -907.891 uV, v_off_fin = -912.764 uV 17:55:18 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 29.645 uVs 17:55:22 I1X Average and RMS of Measurements VT = -28.471 +- 8.643 uVs I1X = 5.694 +- 1.729 uTm 17:55:25 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 17:55:29 Integrated Voltage Measurement... 17:55:38 Voltage Offset Correction: v_off_ini = -902.763 uV, v_off_fin = -934.195 uV 17:55:38 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -157.956 uVs 17:55:38 Integrated Voltage Measurement... 17:55:47 Voltage Offset Correction: v_off_ini = -917.855 uV, v_off_fin = -919.521 uV 17:55:47 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 150.227 uVs 17:55:47 Integrated Voltage Measurement... 17:55:56 Voltage Offset Correction: v_off_ini = -944.292 uV, v_off_fin = -942.235 uV 17:55:56 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -183.402 uVs 17:55:56 Integrated Voltage Measurement... 17:56:05 Voltage Offset Correction: v_off_ini = -944.627 uV, v_off_fin = -935.890 uV 17:56:05 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 164.705 uVs 17:56:05 Integrated Voltage Measurement... 17:56:14 Voltage Offset Correction: v_off_ini = -928.154 uV, v_off_fin = -921.884 uV 17:56:14 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -129.770 uVs 17:56:14 Integrated Voltage Measurement... 17:56:23 Voltage Offset Correction: v_off_ini = -926.730 uV, v_off_fin = -898.909 uV 17:56:23 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5002 mm dXY = -0.5001 mm VT = 174.745 uVs 17:56:27 I1Y Average and RMS of Measurements VT = -160.134 +- 9.871 uVs I1Y = -32.024 +- 1.975 uTm 17:56:32 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 17:56:35 Integrated Voltage Measurement... 17:56:44 Voltage Offset Correction: v_off_ini = -913.376 uV, v_off_fin = -923.711 uV 17:56:44 Integrated Voltage Measurement dXYus = -1.0001 mm, dXYds = 0.0000 mm dXY = -1.0001 mm VT = -60.737 uVs 17:56:44 Integrated Voltage Measurement... 17:56:53 Voltage Offset Correction: v_off_ini = -917.433 uV, v_off_fin = -926.291 uV 17:56:53 Integrated Voltage Measurement dXYus = 0.9998 mm, dXYds = 0.0000 mm dXY = 0.9998 mm VT = 49.271 uVs 17:56:53 Integrated Voltage Measurement... 17:57:02 Voltage Offset Correction: v_off_ini = -921.532 uV, v_off_fin = -924.310 uV 17:57:02 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -30.551 uVs 17:57:02 Integrated Voltage Measurement... 17:57:11 Voltage Offset Correction: v_off_ini = -922.239 uV, v_off_fin = -920.442 uV 17:57:11 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 52.915 uVs 17:57:11 Integrated Voltage Measurement... 17:57:20 Voltage Offset Correction: v_off_ini = -930.232 uV, v_off_fin = -908.931 uV 17:57:20 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -67.630 uVs 17:57:20 Integrated Voltage Measurement... 17:57:29 Voltage Offset Correction: v_off_ini = -936.812 uV, v_off_fin = -933.514 uV 17:57:29 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 61.566 uVs 17:57:31 I2X Average and RMS of Measurements VT = -53.778 +- 9.375 uVs I2X = 4.948 +- 0.862 uTm^2 17:57:36 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 17:57:39 Integrated Voltage Measurement... 17:57:48 Voltage Offset Correction: v_off_ini = -933.147 uV, v_off_fin = -937.669 uV 17:57:48 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -141.626 uVs 17:57:48 Integrated Voltage Measurement... 17:57:57 Voltage Offset Correction: v_off_ini = -936.507 uV, v_off_fin = -912.273 uV 17:57:57 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 139.014 uVs 17:57:57 Integrated Voltage Measurement... 17:58:06 Voltage Offset Correction: v_off_ini = -937.372 uV, v_off_fin = -935.399 uV 17:58:06 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -139.885 uVs 17:58:06 Integrated Voltage Measurement... 17:58:15 Voltage Offset Correction: v_off_ini = -926.392 uV, v_off_fin = -929.072 uV 17:58:15 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 138.092 uVs 17:58:15 Integrated Voltage Measurement... 17:58:24 Voltage Offset Correction: v_off_ini = -927.835 uV, v_off_fin = -924.641 uV 17:58:24 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -176.439 uVs 17:58:24 Integrated Voltage Measurement... 17:58:33 Voltage Offset Correction: v_off_ini = -924.124 uV, v_off_fin = -935.868 uV 17:58:33 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 139.265 uVs 17:58:35 I2Y Average and RMS of Measurements VT = -145.720 +- 8.596 uVs I2Y = -13.406 +- 0.791 uTm^2 17:58:48 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 17:58:53 Integrated Voltage Measurement... 17:59:01 Voltage Offset Correction: v_off_ini = -953.621 uV, v_off_fin = -964.965 uV 17:59:02 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4998 mm dXY = -0.4999 mm VT = -69.335 uVs 17:59:02 Integrated Voltage Measurement... 17:59:10 Voltage Offset Correction: v_off_ini = -951.615 uV, v_off_fin = -942.886 uV 17:59:11 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 48.780 uVs 17:59:11 Integrated Voltage Measurement... 17:59:19 Voltage Offset Correction: v_off_ini = -946.161 uV, v_off_fin = -949.032 uV 17:59:19 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -38.642 uVs 17:59:19 Integrated Voltage Measurement... 17:59:28 Voltage Offset Correction: v_off_ini = -937.345 uV, v_off_fin = -929.054 uV 17:59:28 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 48.058 uVs 17:59:28 Integrated Voltage Measurement... 17:59:37 Voltage Offset Correction: v_off_ini = -942.113 uV, v_off_fin = -943.757 uV 17:59:37 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -33.849 uVs 17:59:37 Integrated Voltage Measurement... 17:59:46 Voltage Offset Correction: v_off_ini = -954.680 uV, v_off_fin = -933.804 uV 17:59:46 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 38.199 uVs 17:59:51 I1X Average and RMS of Measurements VT = -46.144 +- 9.609 uVs I1X = 9.229 +- 1.922 uTm 17:59:53 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 17:59:57 Integrated Voltage Measurement... 18:00:06 Voltage Offset Correction: v_off_ini = -951.724 uV, v_off_fin = -951.763 uV 18:00:06 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -162.777 uVs 18:00:06 Integrated Voltage Measurement... 18:00:15 Voltage Offset Correction: v_off_ini = -933.999 uV, v_off_fin = -981.047 uV 18:00:15 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 165.752 uVs 18:00:15 Integrated Voltage Measurement... 18:00:24 Voltage Offset Correction: v_off_ini = -964.187 uV, v_off_fin = -961.673 uV 18:00:24 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -131.011 uVs 18:00:24 Integrated Voltage Measurement... 18:00:33 Voltage Offset Correction: v_off_ini = -970.319 uV, v_off_fin = -946.179 uV 18:00:33 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 172.315 uVs 18:00:33 Integrated Voltage Measurement... 18:00:42 Voltage Offset Correction: v_off_ini = -936.446 uV, v_off_fin = -965.493 uV 18:00:42 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -144.554 uVs 18:00:42 Integrated Voltage Measurement... 18:00:51 Voltage Offset Correction: v_off_ini = -944.758 uV, v_off_fin = -961.341 uV 18:00:51 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 167.663 uVs 18:00:55 I1Y Average and RMS of Measurements VT = -157.346 +- 5.218 uVs I1Y = -31.467 +- 1.044 uTm 18:01:00 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 18:01:03 Integrated Voltage Measurement... 18:01:12 Voltage Offset Correction: v_off_ini = -962.690 uV, v_off_fin = -968.927 uV 18:01:12 Integrated Voltage Measurement dXYus = -1.0001 mm, dXYds = 0.0000 mm dXY = -1.0001 mm VT = -49.100 uVs 18:01:12 Integrated Voltage Measurement... 18:01:21 Voltage Offset Correction: v_off_ini = -960.100 uV, v_off_fin = -953.888 uV 18:01:21 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 81.325 uVs 18:01:21 Integrated Voltage Measurement... 18:01:30 Voltage Offset Correction: v_off_ini = -956.652 uV, v_off_fin = -937.829 uV 18:01:30 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -57.239 uVs 18:01:30 Integrated Voltage Measurement... 18:01:39 Voltage Offset Correction: v_off_ini = -959.632 uV, v_off_fin = -943.744 uV 18:01:39 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 75.384 uVs 18:01:39 Integrated Voltage Measurement... 18:01:48 Voltage Offset Correction: v_off_ini = -952.355 uV, v_off_fin = -948.725 uV 18:01:48 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -62.181 uVs 18:01:48 Integrated Voltage Measurement... 18:01:57 Voltage Offset Correction: v_off_ini = -955.061 uV, v_off_fin = -947.012 uV 18:01:57 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 42.868 uVs 18:01:59 I2X Average and RMS of Measurements VT = -61.349 +- 6.256 uVs I2X = 5.644 +- 0.575 uTm^2 18:02:04 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 18:02:07 Integrated Voltage Measurement... 18:02:16 Voltage Offset Correction: v_off_ini = -941.931 uV, v_off_fin = -932.420 uV 18:02:16 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -149.346 uVs 18:02:16 Integrated Voltage Measurement... 18:02:25 Voltage Offset Correction: v_off_ini = -959.635 uV, v_off_fin = -929.442 uV 18:02:25 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 163.164 uVs 18:02:25 Integrated Voltage Measurement... 18:02:34 Voltage Offset Correction: v_off_ini = -931.636 uV, v_off_fin = -914.281 uV 18:02:34 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -145.996 uVs 18:02:34 Integrated Voltage Measurement... 18:02:43 Voltage Offset Correction: v_off_ini = -926.424 uV, v_off_fin = -934.153 uV 18:02:43 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 149.982 uVs 18:02:43 Integrated Voltage Measurement... 18:02:52 Voltage Offset Correction: v_off_ini = -913.777 uV, v_off_fin = -925.210 uV 18:02:52 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -163.269 uVs 18:02:52 Integrated Voltage Measurement... 18:03:01 Voltage Offset Correction: v_off_ini = -926.583 uV, v_off_fin = -931.494 uV 18:03:01 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 150.024 uVs 18:03:03 I2Y Average and RMS of Measurements VT = -153.630 +- 3.992 uVs I2Y = -14.134 +- 0.367 uTm^2 18:03:16 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 18:03:21 Integrated Voltage Measurement... 18:03:30 Voltage Offset Correction: v_off_ini = -925.118 uV, v_off_fin = -927.790 uV 18:03:30 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.4998 mm dXY = -0.4999 mm VT = -64.584 uVs 18:03:30 Integrated Voltage Measurement... 18:03:39 Voltage Offset Correction: v_off_ini = -939.661 uV, v_off_fin = -950.189 uV 18:03:39 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 20.483 uVs 18:03:39 Integrated Voltage Measurement... 18:03:48 Voltage Offset Correction: v_off_ini = -931.070 uV, v_off_fin = -949.788 uV 18:03:48 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -60.279 uVs 18:03:48 Integrated Voltage Measurement... 18:03:57 Voltage Offset Correction: v_off_ini = -932.164 uV, v_off_fin = -920.119 uV 18:03:57 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 58.149 uVs 18:03:57 Integrated Voltage Measurement... 18:04:06 Voltage Offset Correction: v_off_ini = -941.034 uV, v_off_fin = -924.957 uV 18:04:06 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -37.363 uVs 18:04:06 Integrated Voltage Measurement... 18:04:15 Voltage Offset Correction: v_off_ini = -957.513 uV, v_off_fin = -945.907 uV 18:04:15 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 46.096 uVs 18:04:19 I1X Average and RMS of Measurements VT = -47.826 +- 8.059 uVs I1X = 9.566 +- 1.611 uTm 18:04:21 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 18:04:26 Integrated Voltage Measurement... 18:04:35 Voltage Offset Correction: v_off_ini = -947.452 uV, v_off_fin = -946.754 uV 18:04:35 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -178.087 uVs 18:04:35 Integrated Voltage Measurement... 18:04:44 Voltage Offset Correction: v_off_ini = -934.436 uV, v_off_fin = -938.532 uV 18:04:44 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = 140.677 uVs 18:04:44 Integrated Voltage Measurement... 18:04:53 Voltage Offset Correction: v_off_ini = -946.059 uV, v_off_fin = -951.216 uV 18:04:53 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -147.427 uVs 18:04:53 Integrated Voltage Measurement... 18:05:01 Voltage Offset Correction: v_off_ini = -946.143 uV, v_off_fin = -935.050 uV 18:05:02 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 151.474 uVs 18:05:02 Integrated Voltage Measurement... 18:05:10 Voltage Offset Correction: v_off_ini = -933.043 uV, v_off_fin = -962.742 uV 18:05:10 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -165.239 uVs 18:05:10 Integrated Voltage Measurement... 18:05:19 Voltage Offset Correction: v_off_ini = -946.767 uV, v_off_fin = -926.814 uV 18:05:19 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 149.754 uVs 18:05:24 I1Y Average and RMS of Measurements VT = -155.443 +- 4.307 uVs I1Y = -31.088 +- 0.862 uTm 18:05:29 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 18:05:31 Integrated Voltage Measurement... 18:05:40 Voltage Offset Correction: v_off_ini = -958.016 uV, v_off_fin = -939.916 uV 18:05:40 Integrated Voltage Measurement dXYus = -1.0002 mm, dXYds = 0.0000 mm dXY = -1.0002 mm VT = -41.496 uVs 18:05:40 Integrated Voltage Measurement... 18:05:49 Voltage Offset Correction: v_off_ini = -939.867 uV, v_off_fin = -916.648 uV 18:05:49 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 25.598 uVs 18:05:49 Integrated Voltage Measurement... 18:05:58 Voltage Offset Correction: v_off_ini = -919.475 uV, v_off_fin = -903.273 uV 18:05:58 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -74.687 uVs 18:05:58 Integrated Voltage Measurement... 18:06:07 Voltage Offset Correction: v_off_ini = -908.413 uV, v_off_fin = -914.324 uV 18:06:07 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 59.148 uVs 18:06:07 Integrated Voltage Measurement... 18:06:16 Voltage Offset Correction: v_off_ini = -910.106 uV, v_off_fin = -914.416 uV 18:06:16 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -100.632 uVs 18:06:16 Integrated Voltage Measurement... 18:06:25 Voltage Offset Correction: v_off_ini = -906.570 uV, v_off_fin = -908.386 uV 18:06:25 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 40.778 uVs 18:06:28 I2X Average and RMS of Measurements VT = -57.056 +- 16.696 uVs I2X = 5.249 +- 1.536 uTm^2 18:06:33 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 18:06:35 Integrated Voltage Measurement... 18:06:44 Voltage Offset Correction: v_off_ini = -911.033 uV, v_off_fin = -919.464 uV 18:06:44 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -159.249 uVs 18:06:44 Integrated Voltage Measurement... 18:06:53 Voltage Offset Correction: v_off_ini = -894.251 uV, v_off_fin = -897.653 uV 18:06:53 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 168.656 uVs 18:06:53 Integrated Voltage Measurement... 18:07:02 Voltage Offset Correction: v_off_ini = -884.234 uV, v_off_fin = -890.832 uV 18:07:02 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -152.413 uVs 18:07:02 Integrated Voltage Measurement... 18:07:11 Voltage Offset Correction: v_off_ini = -895.033 uV, v_off_fin = -920.186 uV 18:07:11 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 167.795 uVs 18:07:11 Integrated Voltage Measurement... 18:07:20 Voltage Offset Correction: v_off_ini = -894.076 uV, v_off_fin = -897.851 uV 18:07:20 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -176.009 uVs 18:07:20 Integrated Voltage Measurement... 18:07:29 Voltage Offset Correction: v_off_ini = -883.908 uV, v_off_fin = -885.902 uV 18:07:29 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 146.745 uVs 18:07:32 I2Y Average and RMS of Measurements VT = -161.811 +- 1.601 uVs I2Y = -14.887 +- 0.147 uTm^2 18:07:34 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 08:51:03 Wire Position: X0 = 50.0000 mm, Y0 = 0.0000 mm