SLAC Magnetic Measurements Date: 01-06-2022 Time: 10:15:13 Project: LCLS-II-HE Magnet Type: Phase Shifter Magnet Name: SXHEPS-000 Dataset Number: 1 Run Purpose: Tuning Run Type: Stretched wire Run Number: 1 Measurement Device: Stretched Wire Operator: YL Comment: 10:15:21 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:15:23 Motion Parameters, X Controller, Axis 1 acc = 0.001000 m/s/s dcl = 0.001000 m/s/s vel = 0.000500 m/s pos = 0.000000 m kp = 188.600000 ki = 820.000000 kd = 491.400000 db = 0.000000 10:15:24 Motion Parameters, X Controller, Axis 2 acc = 0.001000 m/s/s dcl = 0.001000 m/s/s vel = 0.000500 m/s pos = 0.000000 m kp = 188.600000 ki = 820.000000 kd = 491.400000 db = 0.000000 10:15:26 Motion Parameters, Y Controller, Axis 1 acc = 0.001000 m/s/s dcl = 0.001000 m/s/s vel = 0.000500 m/s pos = 0.000000 m kp = 188.600000 ki = 820.000000 kd = 491.400000 db = 0.000000 10:15:28 Motion Parameters, Y Controller, Axis 2 acc = 0.001000 m/s/s dcl = 0.001000 m/s/s vel = 0.000500 m/s pos = 0.000000 m kp = 188.600000 ki = 820.000000 kd = 491.400000 db = 0.000000 10:19:18 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:19:22 Integrated Voltage Measurement... 10:19:32 Voltage Offset Correction: v_off_ini = -170.853 uV, v_off_fin = -175.920 uV 10:19:32 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -68.577 uVs 10:19:32 Integrated Voltage Measurement... 10:19:41 Voltage Offset Correction: v_off_ini = -170.426 uV, v_off_fin = -166.198 uV 10:19:41 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 81.438 uVs 10:19:41 Integrated Voltage Measurement... 10:19:50 Voltage Offset Correction: v_off_ini = -141.698 uV, v_off_fin = -154.170 uV 10:19:50 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -75.342 uVs 10:19:50 Integrated Voltage Measurement... 10:19:59 Voltage Offset Correction: v_off_ini = -139.069 uV, v_off_fin = -139.393 uV 10:19:59 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 64.073 uVs 10:19:59 Integrated Voltage Measurement... 10:20:08 Voltage Offset Correction: v_off_ini = -154.972 uV, v_off_fin = -162.803 uV 10:20:08 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -59.177 uVs 10:20:08 Integrated Voltage Measurement... 10:20:17 Voltage Offset Correction: v_off_ini = -162.290 uV, v_off_fin = -139.923 uV 10:20:17 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = 77.363 uVs 10:20:22 I1X Average and RMS of Measurements VT = -70.995 +- 2.897 uVs I1X = 14.199 +- 0.579 uTm 10:20:24 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:20:28 Integrated Voltage Measurement... 10:20:38 Voltage Offset Correction: v_off_ini = -146.443 uV, v_off_fin = -157.269 uV 10:20:38 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -153.485 uVs 10:20:38 Integrated Voltage Measurement... 10:20:47 Voltage Offset Correction: v_off_ini = -145.942 uV, v_off_fin = -134.884 uV 10:20:47 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 160.792 uVs 10:20:47 Integrated Voltage Measurement... 10:20:56 Voltage Offset Correction: v_off_ini = -147.753 uV, v_off_fin = -164.745 uV 10:20:56 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -187.886 uVs 10:20:56 Integrated Voltage Measurement... 10:21:05 Voltage Offset Correction: v_off_ini = -154.275 uV, v_off_fin = -143.985 uV 10:21:05 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 167.391 uVs 10:21:05 Integrated Voltage Measurement... 10:21:14 Voltage Offset Correction: v_off_ini = -160.239 uV, v_off_fin = -175.056 uV 10:21:14 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -154.750 uVs 10:21:14 Integrated Voltage Measurement... 10:21:23 Voltage Offset Correction: v_off_ini = -149.494 uV, v_off_fin = -137.449 uV 10:21:23 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 159.918 uVs 10:21:28 I1Y Average and RMS of Measurements VT = -164.037 +- 9.618 uVs I1Y = -32.808 +- 1.924 uTm 10:21:33 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:21:36 Integrated Voltage Measurement... 10:21:45 Voltage Offset Correction: v_off_ini = -157.590 uV, v_off_fin = -151.235 uV 10:21:45 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -36.150 uVs 10:21:45 Integrated Voltage Measurement... 10:21:55 Voltage Offset Correction: v_off_ini = -156.162 uV, v_off_fin = -160.483 uV 10:21:55 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 68.102 uVs 10:21:55 Integrated Voltage Measurement... 10:22:04 Voltage Offset Correction: v_off_ini = -149.945 uV, v_off_fin = -161.494 uV 10:22:04 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -48.456 uVs 10:22:04 Integrated Voltage Measurement... 10:22:14 Voltage Offset Correction: v_off_ini = -153.014 uV, v_off_fin = -156.970 uV 10:22:14 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 42.045 uVs 10:22:14 Integrated Voltage Measurement... 10:22:23 Voltage Offset Correction: v_off_ini = -158.785 uV, v_off_fin = -163.346 uV 10:22:23 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -53.153 uVs 10:22:23 Integrated Voltage Measurement... 10:22:33 Voltage Offset Correction: v_off_ini = -167.232 uV, v_off_fin = -160.379 uV 10:22:33 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 40.113 uVs 10:22:35 I2X Average and RMS of Measurements VT = -48.003 +- 2.970 uVs I2X = 3.720 +- 0.230 uTm^2 10:22:40 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:22:43 Integrated Voltage Measurement... 10:22:53 Voltage Offset Correction: v_off_ini = -150.381 uV, v_off_fin = -157.170 uV 10:22:53 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -237.502 uVs 10:22:53 Integrated Voltage Measurement... 10:23:02 Voltage Offset Correction: v_off_ini = -144.789 uV, v_off_fin = -153.231 uV 10:23:02 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 196.018 uVs 10:23:02 Integrated Voltage Measurement... 10:23:12 Voltage Offset Correction: v_off_ini = -152.377 uV, v_off_fin = -161.929 uV 10:23:12 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -206.299 uVs 10:23:12 Integrated Voltage Measurement... 10:23:21 Voltage Offset Correction: v_off_ini = -145.104 uV, v_off_fin = -150.215 uV 10:23:21 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 200.106 uVs 10:23:21 Integrated Voltage Measurement... 10:23:31 Voltage Offset Correction: v_off_ini = -165.312 uV, v_off_fin = -146.568 uV 10:23:31 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -162.956 uVs 10:23:31 Integrated Voltage Measurement... 10:23:40 Voltage Offset Correction: v_off_ini = -152.407 uV, v_off_fin = -128.474 uV 10:23:40 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 180.154 uVs 10:23:43 I2Y Average and RMS of Measurements VT = -197.173 +- 18.941 uVs I2Y = -15.281 +- 1.468 uTm^2 10:23:45 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm