SLAC Magnetic Measurements Date: 02-07-2025 Time: 11:13:35 Project: LCLS-II-HE Magnet Type: Phase Shifter Magnet Name: HE_SXPS_10089 Dataset Number: 1 Run Purpose: Final Results Run Type: Stretched Wire Run Number: 4 Measurement Device: Operator: YL Comment: 11:13:43 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:13:45 Motion Parameters, X Controller, Axis 1 acc = 0.001000 m/s/s dcl = 0.001000 m/s/s vel = 0.001000 m/s pos = 0.000000 m kp = 188.600000 ki = 820.000000 kd = 491.400000 db = 0.000000 11:13:47 Motion Parameters, X Controller, Axis 2 acc = 0.001000 m/s/s dcl = 0.001000 m/s/s vel = 0.001000 m/s pos = 0.000000 m kp = 188.600000 ki = 820.000000 kd = 491.400000 db = 0.000000 11:13:48 Motion Parameters, Y Controller, Axis 1 acc = 0.001000 m/s/s dcl = 0.001000 m/s/s vel = 0.001000 m/s pos = 0.000000 m kp = 188.600000 ki = 820.000000 kd = 491.400000 db = 0.000000 11:13:50 Motion Parameters, Y Controller, Axis 2 acc = 0.001000 m/s/s dcl = 0.001000 m/s/s vel = 0.001000 m/s pos = 0.000000 m kp = 188.600000 ki = 820.000000 kd = 491.400000 db = 0.000000 11:14:25 Wire Position: X0 = -2.0000 mm, Y0 = 0.0000 mm 11:14:30 Integrated Voltage Measurement... 11:14:38 Voltage Offset Correction: v_off_ini = 688.158 uV, v_off_fin = 674.807 uV 11:14:39 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -87.978 uVs 11:14:39 Integrated Voltage Measurement... 11:14:48 Voltage Offset Correction: v_off_ini = 683.003 uV, v_off_fin = 677.081 uV 11:14:48 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 78.899 uVs 11:14:48 Integrated Voltage Measurement... 11:14:57 Voltage Offset Correction: v_off_ini = 685.597 uV, v_off_fin = 685.502 uV 11:14:57 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -86.211 uVs 11:14:57 Integrated Voltage Measurement... 11:15:06 Voltage Offset Correction: v_off_ini = 674.557 uV, v_off_fin = 665.316 uV 11:15:06 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 94.288 uVs 11:15:06 Integrated Voltage Measurement... 11:15:15 Voltage Offset Correction: v_off_ini = 690.393 uV, v_off_fin = 680.262 uV 11:15:15 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -93.110 uVs 11:15:15 Integrated Voltage Measurement... 11:15:24 Voltage Offset Correction: v_off_ini = 675.249 uV, v_off_fin = 657.489 uV 11:15:24 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 73.297 uVs 11:15:28 I1X Average and RMS of Measurements VT = -85.631 +- 3.268 uVs I1X = 17.126 +- 0.653 uTm 11:15:30 Wire Position: X0 = -2.0000 mm, Y0 = 0.0000 mm 11:15:35 Integrated Voltage Measurement... 11:15:44 Voltage Offset Correction: v_off_ini = 650.934 uV, v_off_fin = 656.731 uV 11:15:44 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -227.190 uVs 11:15:44 Integrated Voltage Measurement... 11:15:53 Voltage Offset Correction: v_off_ini = 651.192 uV, v_off_fin = 681.216 uV 11:15:53 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 228.390 uVs 11:15:53 Integrated Voltage Measurement... 11:16:02 Voltage Offset Correction: v_off_ini = 668.815 uV, v_off_fin = 662.096 uV 11:16:02 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -251.959 uVs 11:16:02 Integrated Voltage Measurement... 11:16:11 Voltage Offset Correction: v_off_ini = 686.884 uV, v_off_fin = 693.794 uV 11:16:11 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 236.591 uVs 11:16:11 Integrated Voltage Measurement... 11:16:20 Voltage Offset Correction: v_off_ini = 676.533 uV, v_off_fin = 652.670 uV 11:16:20 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -229.560 uVs 11:16:20 Integrated Voltage Measurement... 11:16:29 Voltage Offset Correction: v_off_ini = 674.392 uV, v_off_fin = 677.280 uV 11:16:29 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.5000 mm dXY = -0.4999 mm VT = 264.414 uVs 11:16:33 I1Y Average and RMS of Measurements VT = -239.684 +- 8.483 uVs I1Y = -47.938 +- 1.697 uTm 11:16:38 Wire Position: X0 = -2.0000 mm, Y0 = 0.0000 mm 11:16:41 Integrated Voltage Measurement... 11:16:50 Voltage Offset Correction: v_off_ini = 645.839 uV, v_off_fin = 638.918 uV 11:16:50 Integrated Voltage Measurement dXYus = -1.0001 mm, dXYds = 0.0000 mm dXY = -1.0001 mm VT = -84.913 uVs 11:16:50 Integrated Voltage Measurement... 11:16:59 Voltage Offset Correction: v_off_ini = 651.542 uV, v_off_fin = 663.927 uV 11:16:59 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 98.571 uVs 11:16:59 Integrated Voltage Measurement... 11:17:08 Voltage Offset Correction: v_off_ini = 648.267 uV, v_off_fin = 648.102 uV 11:17:08 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -82.868 uVs 11:17:08 Integrated Voltage Measurement... 11:17:17 Voltage Offset Correction: v_off_ini = 662.282 uV, v_off_fin = 671.059 uV 11:17:17 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 99.949 uVs 11:17:17 Integrated Voltage Measurement... 11:17:26 Voltage Offset Correction: v_off_ini = 662.681 uV, v_off_fin = 655.293 uV 11:17:26 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -90.723 uVs 11:17:26 Integrated Voltage Measurement... 11:17:35 Voltage Offset Correction: v_off_ini = 660.600 uV, v_off_fin = 671.780 uV 11:17:35 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 109.043 uVs 11:17:37 I2X Average and RMS of Measurements VT = -94.344 +- 3.919 uVs I2X = 8.680 +- 0.361 uTm^2 11:17:42 Wire Position: X0 = -2.0000 mm, Y0 = 0.0000 mm 11:17:45 Integrated Voltage Measurement... 11:17:54 Voltage Offset Correction: v_off_ini = 650.395 uV, v_off_fin = 648.916 uV 11:17:54 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -339.344 uVs 11:17:54 Integrated Voltage Measurement... 11:18:03 Voltage Offset Correction: v_off_ini = 636.406 uV, v_off_fin = 654.385 uV 11:18:03 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 352.080 uVs 11:18:03 Integrated Voltage Measurement... 11:18:12 Voltage Offset Correction: v_off_ini = 667.872 uV, v_off_fin = 659.483 uV 11:18:12 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -318.990 uVs 11:18:12 Integrated Voltage Measurement... 11:18:21 Voltage Offset Correction: v_off_ini = 651.588 uV, v_off_fin = 643.177 uV 11:18:21 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 339.641 uVs 11:18:21 Integrated Voltage Measurement... 11:18:30 Voltage Offset Correction: v_off_ini = 611.360 uV, v_off_fin = 636.941 uV 11:18:30 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -330.116 uVs 11:18:30 Integrated Voltage Measurement... 11:18:39 Voltage Offset Correction: v_off_ini = 657.503 uV, v_off_fin = 643.598 uV 11:18:39 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 336.401 uVs 11:18:42 I2Y Average and RMS of Measurements VT = -336.095 +- 6.988 uVs I2Y = -30.921 +- 0.643 uTm^2 11:18:46 Wire Position: X0 = -1.0000 mm, Y0 = 0.0000 mm 11:18:50 Integrated Voltage Measurement... 11:18:59 Voltage Offset Correction: v_off_ini = 648.922 uV, v_off_fin = 638.743 uV 11:18:59 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.4998 mm dXY = -0.4999 mm VT = -92.802 uVs 11:19:00 Integrated Voltage Measurement... 11:19:08 Voltage Offset Correction: v_off_ini = 649.575 uV, v_off_fin = 635.914 uV 11:19:09 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 98.510 uVs 11:19:09 Integrated Voltage Measurement... 11:19:18 Voltage Offset Correction: v_off_ini = 661.230 uV, v_off_fin = 647.300 uV 11:19:18 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -78.858 uVs 11:19:18 Integrated Voltage Measurement... 11:19:27 Voltage Offset Correction: v_off_ini = 650.764 uV, v_off_fin = 644.720 uV 11:19:27 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 104.092 uVs 11:19:27 Integrated Voltage Measurement... 11:19:36 Voltage Offset Correction: v_off_ini = 657.779 uV, v_off_fin = 660.710 uV 11:19:36 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -79.324 uVs 11:19:36 Integrated Voltage Measurement... 11:19:45 Voltage Offset Correction: v_off_ini = 656.245 uV, v_off_fin = 661.946 uV 11:19:45 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 79.604 uVs 11:19:49 I1X Average and RMS of Measurements VT = -88.865 +- 6.863 uVs I1X = 17.774 +- 1.374 uTm 11:19:51 Wire Position: X0 = -1.0000 mm, Y0 = 0.0000 mm 11:19:56 Integrated Voltage Measurement... 11:20:05 Voltage Offset Correction: v_off_ini = 656.219 uV, v_off_fin = 630.091 uV 11:20:05 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -263.330 uVs 11:20:05 Integrated Voltage Measurement... 11:20:14 Voltage Offset Correction: v_off_ini = 641.251 uV, v_off_fin = 666.390 uV 11:20:14 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 258.414 uVs 11:20:14 Integrated Voltage Measurement... 11:20:23 Voltage Offset Correction: v_off_ini = 655.473 uV, v_off_fin = 633.555 uV 11:20:23 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -230.469 uVs 11:20:23 Integrated Voltage Measurement... 11:20:32 Voltage Offset Correction: v_off_ini = 657.332 uV, v_off_fin = 682.131 uV 11:20:32 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 227.654 uVs 11:20:32 Integrated Voltage Measurement... 11:20:41 Voltage Offset Correction: v_off_ini = 661.386 uV, v_off_fin = 634.796 uV 11:20:41 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -236.516 uVs 11:20:41 Integrated Voltage Measurement... 11:20:50 Voltage Offset Correction: v_off_ini = 636.207 uV, v_off_fin = 658.369 uV 11:20:50 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 251.256 uVs 11:20:54 I1Y Average and RMS of Measurements VT = -244.607 +- 12.997 uVs I1Y = -48.921 +- 2.599 uTm 11:20:59 Wire Position: X0 = -1.0000 mm, Y0 = 0.0000 mm 11:21:02 Integrated Voltage Measurement... 11:21:11 Voltage Offset Correction: v_off_ini = 636.426 uV, v_off_fin = 628.929 uV 11:21:11 Integrated Voltage Measurement dXYus = -1.0001 mm, dXYds = 0.0000 mm dXY = -1.0001 mm VT = -96.415 uVs 11:21:11 Integrated Voltage Measurement... 11:21:20 Voltage Offset Correction: v_off_ini = 635.061 uV, v_off_fin = 641.402 uV 11:21:20 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 87.802 uVs 11:21:20 Integrated Voltage Measurement... 11:21:29 Voltage Offset Correction: v_off_ini = 647.691 uV, v_off_fin = 651.313 uV 11:21:29 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -65.807 uVs 11:21:29 Integrated Voltage Measurement... 11:21:38 Voltage Offset Correction: v_off_ini = 647.152 uV, v_off_fin = 668.746 uV 11:21:38 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 92.179 uVs 11:21:38 Integrated Voltage Measurement... 11:21:47 Voltage Offset Correction: v_off_ini = 647.646 uV, v_off_fin = 635.727 uV 11:21:47 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -92.067 uVs 11:21:47 Integrated Voltage Measurement... 11:21:56 Voltage Offset Correction: v_off_ini = 645.198 uV, v_off_fin = 650.214 uV 11:21:56 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 91.087 uVs 11:21:58 I2X Average and RMS of Measurements VT = -87.559 +- 6.061 uVs I2X = 8.055 +- 0.558 uTm^2 11:22:03 Wire Position: X0 = -1.0000 mm, Y0 = 0.0000 mm 11:22:06 Integrated Voltage Measurement... 11:22:15 Voltage Offset Correction: v_off_ini = 640.539 uV, v_off_fin = 656.817 uV 11:22:15 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -328.387 uVs 11:22:15 Integrated Voltage Measurement... 11:22:24 Voltage Offset Correction: v_off_ini = 659.358 uV, v_off_fin = 652.013 uV 11:22:24 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 327.274 uVs 11:22:24 Integrated Voltage Measurement... 11:22:33 Voltage Offset Correction: v_off_ini = 639.789 uV, v_off_fin = 654.527 uV 11:22:33 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -333.787 uVs 11:22:33 Integrated Voltage Measurement... 11:22:42 Voltage Offset Correction: v_off_ini = 655.307 uV, v_off_fin = 660.480 uV 11:22:42 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 334.107 uVs 11:22:42 Integrated Voltage Measurement... 11:22:51 Voltage Offset Correction: v_off_ini = 642.487 uV, v_off_fin = 649.642 uV 11:22:51 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -354.468 uVs 11:22:51 Integrated Voltage Measurement... 11:23:00 Voltage Offset Correction: v_off_ini = 650.129 uV, v_off_fin = 655.398 uV 11:23:00 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 321.602 uVs 11:23:03 I2Y Average and RMS of Measurements VT = -333.271 +- 4.193 uVs I2Y = -30.661 +- 0.386 uTm^2 11:23:07 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:23:12 Integrated Voltage Measurement... 11:23:21 Voltage Offset Correction: v_off_ini = 676.581 uV, v_off_fin = 664.779 uV 11:23:21 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -94.312 uVs 11:23:21 Integrated Voltage Measurement... 11:23:30 Voltage Offset Correction: v_off_ini = 674.721 uV, v_off_fin = 674.396 uV 11:23:30 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 99.785 uVs 11:23:30 Integrated Voltage Measurement... 11:23:39 Voltage Offset Correction: v_off_ini = 654.423 uV, v_off_fin = 668.335 uV 11:23:39 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -71.662 uVs 11:23:39 Integrated Voltage Measurement... 11:23:48 Voltage Offset Correction: v_off_ini = 659.426 uV, v_off_fin = 664.885 uV 11:23:48 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 88.927 uVs 11:23:48 Integrated Voltage Measurement... 11:23:57 Voltage Offset Correction: v_off_ini = 666.738 uV, v_off_fin = 678.107 uV 11:23:57 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -64.664 uVs 11:23:57 Integrated Voltage Measurement... 11:24:06 Voltage Offset Correction: v_off_ini = 669.814 uV, v_off_fin = 670.735 uV 11:24:06 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 72.937 uVs 11:24:10 I1X Average and RMS of Measurements VT = -82.048 +- 11.599 uVs I1X = 16.410 +- 2.320 uTm 11:24:12 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:24:17 Integrated Voltage Measurement... 11:24:26 Voltage Offset Correction: v_off_ini = 667.815 uV, v_off_fin = 641.386 uV 11:24:26 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = -248.828 uVs 11:24:26 Integrated Voltage Measurement... 11:24:35 Voltage Offset Correction: v_off_ini = 666.418 uV, v_off_fin = 682.779 uV 11:24:35 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 235.421 uVs 11:24:35 Integrated Voltage Measurement... 11:24:44 Voltage Offset Correction: v_off_ini = 655.050 uV, v_off_fin = 652.454 uV 11:24:44 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -233.513 uVs 11:24:44 Integrated Voltage Measurement... 11:24:53 Voltage Offset Correction: v_off_ini = 667.986 uV, v_off_fin = 671.441 uV 11:24:53 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 241.919 uVs 11:24:53 Integrated Voltage Measurement... 11:25:02 Voltage Offset Correction: v_off_ini = 676.967 uV, v_off_fin = 653.465 uV 11:25:02 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -248.893 uVs 11:25:02 Integrated Voltage Measurement... 11:25:11 Voltage Offset Correction: v_off_ini = 670.246 uV, v_off_fin = 671.735 uV 11:25:11 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 234.532 uVs 11:25:15 I1Y Average and RMS of Measurements VT = -240.518 +- 1.988 uVs I1Y = -48.104 +- 0.398 uTm 11:25:20 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:25:23 Integrated Voltage Measurement... 11:25:32 Voltage Offset Correction: v_off_ini = 662.631 uV, v_off_fin = 668.574 uV 11:25:32 Integrated Voltage Measurement dXYus = -1.0001 mm, dXYds = 0.0000 mm dXY = -1.0001 mm VT = -87.885 uVs 11:25:32 Integrated Voltage Measurement... 11:25:41 Voltage Offset Correction: v_off_ini = 681.936 uV, v_off_fin = 686.369 uV 11:25:41 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 101.182 uVs 11:25:41 Integrated Voltage Measurement... 11:25:50 Voltage Offset Correction: v_off_ini = 690.533 uV, v_off_fin = 682.638 uV 11:25:50 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -103.517 uVs 11:25:50 Integrated Voltage Measurement... 11:25:59 Voltage Offset Correction: v_off_ini = 696.084 uV, v_off_fin = 706.511 uV 11:25:59 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 76.896 uVs 11:25:59 Integrated Voltage Measurement... 11:26:08 Voltage Offset Correction: v_off_ini = 678.379 uV, v_off_fin = 688.480 uV 11:26:08 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -82.955 uVs 11:26:08 Integrated Voltage Measurement... 11:26:17 Voltage Offset Correction: v_off_ini = 697.604 uV, v_off_fin = 684.670 uV 11:26:17 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 80.823 uVs 11:26:19 I2X Average and RMS of Measurements VT = -88.876 +- 5.247 uVs I2X = 8.177 +- 0.483 uTm^2 11:26:24 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:26:27 Integrated Voltage Measurement... 11:26:36 Voltage Offset Correction: v_off_ini = 684.328 uV, v_off_fin = 700.100 uV 11:26:36 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -325.216 uVs 11:26:36 Integrated Voltage Measurement... 11:26:45 Voltage Offset Correction: v_off_ini = 683.875 uV, v_off_fin = 692.306 uV 11:26:45 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 340.498 uVs 11:26:45 Integrated Voltage Measurement... 11:26:54 Voltage Offset Correction: v_off_ini = 688.740 uV, v_off_fin = 666.543 uV 11:26:54 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -337.641 uVs 11:26:54 Integrated Voltage Measurement... 11:27:03 Voltage Offset Correction: v_off_ini = 665.001 uV, v_off_fin = 691.110 uV 11:27:03 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 341.610 uVs 11:27:03 Integrated Voltage Measurement... 11:27:12 Voltage Offset Correction: v_off_ini = 673.654 uV, v_off_fin = 676.591 uV 11:27:12 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -300.615 uVs 11:27:12 Integrated Voltage Measurement... 11:27:21 Voltage Offset Correction: v_off_ini = 682.904 uV, v_off_fin = 680.643 uV 11:27:21 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 336.939 uVs 11:27:24 I2Y Average and RMS of Measurements VT = -330.420 +- 8.684 uVs I2Y = -30.399 +- 0.799 uTm^2 11:27:28 Wire Position: X0 = 1.0000 mm, Y0 = 0.0000 mm 11:27:33 Integrated Voltage Measurement... 11:27:42 Voltage Offset Correction: v_off_ini = 691.422 uV, v_off_fin = 704.294 uV 11:27:42 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -75.357 uVs 11:27:42 Integrated Voltage Measurement... 11:27:51 Voltage Offset Correction: v_off_ini = 711.127 uV, v_off_fin = 677.579 uV 11:27:51 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 72.041 uVs 11:27:51 Integrated Voltage Measurement... 11:28:00 Voltage Offset Correction: v_off_ini = 686.680 uV, v_off_fin = 697.777 uV 11:28:00 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -65.499 uVs 11:28:00 Integrated Voltage Measurement... 11:28:09 Voltage Offset Correction: v_off_ini = 719.590 uV, v_off_fin = 709.505 uV 11:28:09 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 79.180 uVs 11:28:09 Integrated Voltage Measurement... 11:28:18 Voltage Offset Correction: v_off_ini = 726.079 uV, v_off_fin = 708.054 uV 11:28:18 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -88.602 uVs 11:28:18 Integrated Voltage Measurement... 11:28:27 Voltage Offset Correction: v_off_ini = 707.440 uV, v_off_fin = 700.688 uV 11:28:27 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 71.947 uVs 11:28:31 I1X Average and RMS of Measurements VT = -75.438 +- 3.465 uVs I1X = 15.088 +- 0.693 uTm 11:28:34 Wire Position: X0 = 1.0000 mm, Y0 = 0.0000 mm 11:28:38 Integrated Voltage Measurement... 11:28:47 Voltage Offset Correction: v_off_ini = 715.473 uV, v_off_fin = 712.306 uV 11:28:47 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -262.508 uVs 11:28:47 Integrated Voltage Measurement... 11:28:56 Voltage Offset Correction: v_off_ini = 707.917 uV, v_off_fin = 714.550 uV 11:28:56 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 254.253 uVs 11:28:56 Integrated Voltage Measurement... 11:29:05 Voltage Offset Correction: v_off_ini = 696.661 uV, v_off_fin = 684.623 uV 11:29:05 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -237.338 uVs 11:29:05 Integrated Voltage Measurement... 11:29:14 Voltage Offset Correction: v_off_ini = 680.826 uV, v_off_fin = 689.181 uV 11:29:14 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 259.536 uVs 11:29:14 Integrated Voltage Measurement... 11:29:23 Voltage Offset Correction: v_off_ini = 699.022 uV, v_off_fin = 683.213 uV 11:29:23 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -242.961 uVs 11:29:23 Integrated Voltage Measurement... 11:29:32 Voltage Offset Correction: v_off_ini = 676.799 uV, v_off_fin = 687.840 uV 11:29:32 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 242.131 uVs 11:29:36 I1Y Average and RMS of Measurements VT = -249.788 +- 6.534 uVs I1Y = -49.958 +- 1.307 uTm 11:29:41 Wire Position: X0 = 1.0000 mm, Y0 = 0.0000 mm 11:29:44 Integrated Voltage Measurement... 11:29:53 Voltage Offset Correction: v_off_ini = 691.325 uV, v_off_fin = 682.418 uV 11:29:53 Integrated Voltage Measurement dXYus = -1.0001 mm, dXYds = 0.0000 mm dXY = -1.0001 mm VT = -85.329 uVs 11:29:53 Integrated Voltage Measurement... 11:30:02 Voltage Offset Correction: v_off_ini = 689.545 uV, v_off_fin = 683.048 uV 11:30:02 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 67.336 uVs 11:30:02 Integrated Voltage Measurement... 11:30:11 Voltage Offset Correction: v_off_ini = 667.611 uV, v_off_fin = 665.366 uV 11:30:11 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -74.623 uVs 11:30:11 Integrated Voltage Measurement... 11:30:20 Voltage Offset Correction: v_off_ini = 697.758 uV, v_off_fin = 696.262 uV 11:30:20 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 81.729 uVs 11:30:20 Integrated Voltage Measurement... 11:30:29 Voltage Offset Correction: v_off_ini = 679.727 uV, v_off_fin = 660.500 uV 11:30:29 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -104.077 uVs 11:30:29 Integrated Voltage Measurement... 11:30:38 Voltage Offset Correction: v_off_ini = 678.531 uV, v_off_fin = 696.162 uV 11:30:38 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 81.713 uVs 11:30:40 I2X Average and RMS of Measurements VT = -82.468 +- 7.411 uVs I2X = 7.587 +- 0.682 uTm^2 11:30:45 Wire Position: X0 = 1.0000 mm, Y0 = 0.0000 mm 11:30:48 Integrated Voltage Measurement... 11:30:57 Voltage Offset Correction: v_off_ini = 673.748 uV, v_off_fin = 668.990 uV 11:30:57 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = -312.469 uVs 11:30:57 Integrated Voltage Measurement... 11:31:06 Voltage Offset Correction: v_off_ini = 678.512 uV, v_off_fin = 674.492 uV 11:31:06 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 329.073 uVs 11:31:06 Integrated Voltage Measurement... 11:31:15 Voltage Offset Correction: v_off_ini = 667.546 uV, v_off_fin = 697.429 uV 11:31:15 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -345.795 uVs 11:31:15 Integrated Voltage Measurement... 11:31:24 Voltage Offset Correction: v_off_ini = 699.945 uV, v_off_fin = 708.331 uV 11:31:24 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 343.403 uVs 11:31:24 Integrated Voltage Measurement... 11:31:33 Voltage Offset Correction: v_off_ini = 704.536 uV, v_off_fin = 712.123 uV 11:31:33 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -335.789 uVs 11:31:33 Integrated Voltage Measurement... 11:31:42 Voltage Offset Correction: v_off_ini = 699.034 uV, v_off_fin = 682.727 uV 11:31:42 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 316.271 uVs 11:31:45 I2Y Average and RMS of Measurements VT = -330.466 +- 10.221 uVs I2Y = -30.403 +- 0.940 uTm^2 11:31:49 Wire Position: X0 = 2.0000 mm, Y0 = 0.0000 mm 11:31:54 Integrated Voltage Measurement... 11:32:03 Voltage Offset Correction: v_off_ini = 677.830 uV, v_off_fin = 659.805 uV 11:32:03 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.4998 mm dXY = -0.4999 mm VT = -77.756 uVs 11:32:03 Integrated Voltage Measurement... 11:32:12 Voltage Offset Correction: v_off_ini = 658.399 uV, v_off_fin = 658.832 uV 11:32:12 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 77.036 uVs 11:32:12 Integrated Voltage Measurement... 11:32:21 Voltage Offset Correction: v_off_ini = 658.978 uV, v_off_fin = 656.155 uV 11:32:21 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -66.171 uVs 11:32:21 Integrated Voltage Measurement... 11:32:30 Voltage Offset Correction: v_off_ini = 666.933 uV, v_off_fin = 659.576 uV 11:32:30 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 67.052 uVs 11:32:30 Integrated Voltage Measurement... 11:32:39 Voltage Offset Correction: v_off_ini = 678.557 uV, v_off_fin = 651.409 uV 11:32:39 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -82.896 uVs 11:32:39 Integrated Voltage Measurement... 11:32:48 Voltage Offset Correction: v_off_ini = 646.904 uV, v_off_fin = 668.832 uV 11:32:48 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 65.184 uVs 11:32:52 I1X Average and RMS of Measurements VT = -72.682 +- 4.506 uVs I1X = 14.537 +- 0.902 uTm 11:32:54 Wire Position: X0 = 2.0000 mm, Y0 = 0.0000 mm 11:32:59 Integrated Voltage Measurement... 11:33:08 Voltage Offset Correction: v_off_ini = 674.736 uV, v_off_fin = 657.602 uV 11:33:08 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -244.855 uVs 11:33:08 Integrated Voltage Measurement... 11:33:17 Voltage Offset Correction: v_off_ini = 674.621 uV, v_off_fin = 685.231 uV 11:33:17 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 252.812 uVs 11:33:17 Integrated Voltage Measurement... 11:33:26 Voltage Offset Correction: v_off_ini = 673.002 uV, v_off_fin = 671.003 uV 11:33:26 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -248.126 uVs 11:33:26 Integrated Voltage Measurement... 11:33:35 Voltage Offset Correction: v_off_ini = 678.496 uV, v_off_fin = 693.937 uV 11:33:35 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 252.108 uVs 11:33:35 Integrated Voltage Measurement... 11:33:44 Voltage Offset Correction: v_off_ini = 696.679 uV, v_off_fin = 690.688 uV 11:33:44 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -234.692 uVs 11:33:44 Integrated Voltage Measurement... 11:33:53 Voltage Offset Correction: v_off_ini = 698.011 uV, v_off_fin = 703.276 uV 11:33:53 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 253.046 uVs 11:33:57 I1Y Average and RMS of Measurements VT = -247.607 +- 2.694 uVs I1Y = -49.521 +- 0.539 uTm 11:34:02 Wire Position: X0 = 2.0000 mm, Y0 = 0.0000 mm 11:34:05 Integrated Voltage Measurement... 11:34:13 Voltage Offset Correction: v_off_ini = 666.913 uV, v_off_fin = 668.032 uV 11:34:14 Integrated Voltage Measurement dXYus = -1.0002 mm, dXYds = 0.0000 mm dXY = -1.0002 mm VT = -78.343 uVs 11:34:14 Integrated Voltage Measurement... 11:34:23 Voltage Offset Correction: v_off_ini = 677.896 uV, v_off_fin = 695.078 uV 11:34:23 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 67.272 uVs 11:34:23 Integrated Voltage Measurement... 11:34:32 Voltage Offset Correction: v_off_ini = 677.289 uV, v_off_fin = 668.731 uV 11:34:32 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -95.745 uVs 11:34:32 Integrated Voltage Measurement... 11:34:41 Voltage Offset Correction: v_off_ini = 684.182 uV, v_off_fin = 699.771 uV 11:34:41 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 72.127 uVs 11:34:41 Integrated Voltage Measurement... 11:34:50 Voltage Offset Correction: v_off_ini = 690.876 uV, v_off_fin = 701.927 uV 11:34:50 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -81.513 uVs 11:34:50 Integrated Voltage Measurement... 11:34:59 Voltage Offset Correction: v_off_ini = 691.402 uV, v_off_fin = 706.995 uV 11:34:59 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 80.163 uVs 11:35:01 I2X Average and RMS of Measurements VT = -79.194 +- 4.690 uVs I2X = 7.286 +- 0.432 uTm^2 11:35:06 Wire Position: X0 = 2.0000 mm, Y0 = 0.0000 mm 11:35:09 Integrated Voltage Measurement... 11:35:18 Voltage Offset Correction: v_off_ini = 694.510 uV, v_off_fin = 698.682 uV 11:35:18 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -318.132 uVs 11:35:18 Integrated Voltage Measurement... 11:35:27 Voltage Offset Correction: v_off_ini = 705.459 uV, v_off_fin = 705.452 uV 11:35:27 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 342.121 uVs 11:35:27 Integrated Voltage Measurement... 11:35:36 Voltage Offset Correction: v_off_ini = 701.388 uV, v_off_fin = 691.954 uV 11:35:36 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -364.447 uVs 11:35:36 Integrated Voltage Measurement... 11:35:45 Voltage Offset Correction: v_off_ini = 707.093 uV, v_off_fin = 682.911 uV 11:35:45 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 365.440 uVs 11:35:45 Integrated Voltage Measurement... 11:35:54 Voltage Offset Correction: v_off_ini = 689.036 uV, v_off_fin = 694.313 uV 11:35:54 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -363.178 uVs 11:35:54 Integrated Voltage Measurement... 11:36:03 Voltage Offset Correction: v_off_ini = 697.643 uV, v_off_fin = 714.015 uV 11:36:03 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 336.733 uVs 11:36:05 I2Y Average and RMS of Measurements VT = -348.342 +- 14.260 uVs I2Y = -32.047 +- 1.312 uTm^2 11:36:13 Wire Position: X0 = 0.0000 mm, Y0 = -1.0000 mm 11:36:17 Integrated Voltage Measurement... 11:36:26 Voltage Offset Correction: v_off_ini = 713.528 uV, v_off_fin = 680.225 uV 11:36:26 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.4998 mm dXY = -0.4999 mm VT = -81.664 uVs 11:36:26 Integrated Voltage Measurement... 11:36:35 Voltage Offset Correction: v_off_ini = 696.710 uV, v_off_fin = 687.839 uV 11:36:35 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 78.701 uVs 11:36:35 Integrated Voltage Measurement... 11:36:44 Voltage Offset Correction: v_off_ini = 695.695 uV, v_off_fin = 708.135 uV 11:36:44 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -65.232 uVs 11:36:44 Integrated Voltage Measurement... 11:36:53 Voltage Offset Correction: v_off_ini = 707.482 uV, v_off_fin = 698.789 uV 11:36:53 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 90.501 uVs 11:36:53 Integrated Voltage Measurement... 11:37:02 Voltage Offset Correction: v_off_ini = 711.747 uV, v_off_fin = 705.214 uV 11:37:02 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -73.830 uVs 11:37:02 Integrated Voltage Measurement... 11:37:11 Voltage Offset Correction: v_off_ini = 687.478 uV, v_off_fin = 693.428 uV 11:37:11 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 80.520 uVs 11:37:16 I1X Average and RMS of Measurements VT = -78.408 +- 1.286 uVs I1X = 15.682 +- 0.258 uTm 11:37:18 Wire Position: X0 = 0.0000 mm, Y0 = -1.0000 mm 11:37:22 Integrated Voltage Measurement... 11:37:31 Voltage Offset Correction: v_off_ini = 691.453 uV, v_off_fin = 691.792 uV 11:37:31 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -243.219 uVs 11:37:31 Integrated Voltage Measurement... 11:37:40 Voltage Offset Correction: v_off_ini = 708.186 uV, v_off_fin = 724.780 uV 11:37:40 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 239.633 uVs 11:37:40 Integrated Voltage Measurement... 11:37:49 Voltage Offset Correction: v_off_ini = 722.399 uV, v_off_fin = 714.425 uV 11:37:49 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -263.968 uVs 11:37:49 Integrated Voltage Measurement... 11:37:58 Voltage Offset Correction: v_off_ini = 727.259 uV, v_off_fin = 737.107 uV 11:37:58 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 244.428 uVs 11:37:58 Integrated Voltage Measurement... 11:38:07 Voltage Offset Correction: v_off_ini = 703.804 uV, v_off_fin = 688.038 uV 11:38:07 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -235.574 uVs 11:38:07 Integrated Voltage Measurement... 11:38:16 Voltage Offset Correction: v_off_ini = 704.576 uV, v_off_fin = 724.201 uV 11:38:16 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 247.239 uVs 11:38:20 I1Y Average and RMS of Measurements VT = -245.677 +- 6.025 uVs I1Y = -49.135 +- 1.205 uTm 11:38:23 Wire Position: X0 = 0.0000 mm, Y0 = -1.0000 mm 11:38:26 Integrated Voltage Measurement... 11:38:35 Voltage Offset Correction: v_off_ini = 723.340 uV, v_off_fin = 712.866 uV 11:38:35 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -104.121 uVs 11:38:35 Integrated Voltage Measurement... 11:38:44 Voltage Offset Correction: v_off_ini = 724.944 uV, v_off_fin = 718.922 uV 11:38:44 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 109.429 uVs 11:38:44 Integrated Voltage Measurement... 11:38:53 Voltage Offset Correction: v_off_ini = 689.873 uV, v_off_fin = 703.685 uV 11:38:53 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -114.928 uVs 11:38:53 Integrated Voltage Measurement... 11:39:02 Voltage Offset Correction: v_off_ini = 701.930 uV, v_off_fin = 729.798 uV 11:39:02 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 124.035 uVs 11:39:02 Integrated Voltage Measurement... 11:39:11 Voltage Offset Correction: v_off_ini = 704.530 uV, v_off_fin = 711.503 uV 11:39:11 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -106.876 uVs 11:39:11 Integrated Voltage Measurement... 11:39:20 Voltage Offset Correction: v_off_ini = 728.490 uV, v_off_fin = 752.789 uV 11:39:20 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 113.638 uVs 11:39:22 I2X Average and RMS of Measurements VT = -112.171 +- 5.361 uVs I2X = 10.320 +- 0.493 uTm^2 11:39:25 Wire Position: X0 = 0.0000 mm, Y0 = -1.0000 mm 11:39:27 Integrated Voltage Measurement... 11:39:36 Voltage Offset Correction: v_off_ini = 728.842 uV, v_off_fin = 719.482 uV 11:39:36 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -336.379 uVs 11:39:36 Integrated Voltage Measurement... 11:39:45 Voltage Offset Correction: v_off_ini = 737.241 uV, v_off_fin = 725.113 uV 11:39:45 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 324.699 uVs 11:39:45 Integrated Voltage Measurement... 11:39:54 Voltage Offset Correction: v_off_ini = 727.750 uV, v_off_fin = 734.590 uV 11:39:54 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -317.471 uVs 11:39:54 Integrated Voltage Measurement... 11:40:03 Voltage Offset Correction: v_off_ini = 726.793 uV, v_off_fin = 738.727 uV 11:40:03 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 340.718 uVs 11:40:03 Integrated Voltage Measurement... 11:40:12 Voltage Offset Correction: v_off_ini = 753.979 uV, v_off_fin = 738.668 uV 11:40:12 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -342.192 uVs 11:40:12 Integrated Voltage Measurement... 11:40:21 Voltage Offset Correction: v_off_ini = 737.541 uV, v_off_fin = 740.982 uV 11:40:21 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 330.257 uVs 11:40:24 I2Y Average and RMS of Measurements VT = -331.953 +- 3.078 uVs I2Y = -30.540 +- 0.283 uTm^2 11:40:28 Wire Position: X0 = 0.0000 mm, Y0 = -0.5000 mm 11:40:32 Integrated Voltage Measurement... 11:40:41 Voltage Offset Correction: v_off_ini = 753.095 uV, v_off_fin = 750.257 uV 11:40:41 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -71.120 uVs 11:40:41 Integrated Voltage Measurement... 11:40:50 Voltage Offset Correction: v_off_ini = 735.822 uV, v_off_fin = 735.144 uV 11:40:50 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 64.306 uVs 11:40:50 Integrated Voltage Measurement... 11:40:59 Voltage Offset Correction: v_off_ini = 755.110 uV, v_off_fin = 751.225 uV 11:40:59 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -93.351 uVs 11:40:59 Integrated Voltage Measurement... 11:41:08 Voltage Offset Correction: v_off_ini = 725.982 uV, v_off_fin = 734.424 uV 11:41:08 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 90.681 uVs 11:41:08 Integrated Voltage Measurement... 11:41:17 Voltage Offset Correction: v_off_ini = 739.481 uV, v_off_fin = 732.853 uV 11:41:17 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -96.620 uVs 11:41:17 Integrated Voltage Measurement... 11:41:26 Voltage Offset Correction: v_off_ini = 734.485 uV, v_off_fin = 745.015 uV 11:41:26 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 60.404 uVs 11:41:30 I1X Average and RMS of Measurements VT = -79.414 +- 9.942 uVs I1X = 15.883 +- 1.988 uTm 11:41:33 Wire Position: X0 = 0.0000 mm, Y0 = -0.5000 mm 11:41:37 Integrated Voltage Measurement... 11:41:46 Voltage Offset Correction: v_off_ini = 747.133 uV, v_off_fin = 737.815 uV 11:41:46 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -231.156 uVs 11:41:46 Integrated Voltage Measurement... 11:41:55 Voltage Offset Correction: v_off_ini = 732.331 uV, v_off_fin = 749.758 uV 11:41:55 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 239.980 uVs 11:41:55 Integrated Voltage Measurement... 11:42:04 Voltage Offset Correction: v_off_ini = 733.422 uV, v_off_fin = 717.146 uV 11:42:04 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -242.123 uVs 11:42:04 Integrated Voltage Measurement... 11:42:13 Voltage Offset Correction: v_off_ini = 714.897 uV, v_off_fin = 739.186 uV 11:42:13 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 240.018 uVs 11:42:13 Integrated Voltage Measurement... 11:42:22 Voltage Offset Correction: v_off_ini = 713.671 uV, v_off_fin = 723.070 uV 11:42:22 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -247.567 uVs 11:42:22 Integrated Voltage Measurement... 11:42:31 Voltage Offset Correction: v_off_ini = 727.814 uV, v_off_fin = 731.682 uV 11:42:31 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 233.320 uVs 11:42:35 I1Y Average and RMS of Measurements VT = -239.027 +- 2.460 uVs I1Y = -47.805 +- 0.492 uTm 11:42:38 Wire Position: X0 = 0.0000 mm, Y0 = -0.5000 mm 11:42:40 Integrated Voltage Measurement... 11:42:49 Voltage Offset Correction: v_off_ini = 727.690 uV, v_off_fin = 728.536 uV 11:42:49 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -103.611 uVs 11:42:49 Integrated Voltage Measurement... 11:42:58 Voltage Offset Correction: v_off_ini = 741.361 uV, v_off_fin = 734.986 uV 11:42:58 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 132.456 uVs 11:42:58 Integrated Voltage Measurement... 11:43:07 Voltage Offset Correction: v_off_ini = 726.749 uV, v_off_fin = 722.613 uV 11:43:07 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -98.420 uVs 11:43:07 Integrated Voltage Measurement... 11:43:16 Voltage Offset Correction: v_off_ini = 713.838 uV, v_off_fin = 723.830 uV 11:43:16 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 107.812 uVs 11:43:16 Integrated Voltage Measurement... 11:43:25 Voltage Offset Correction: v_off_ini = 702.529 uV, v_off_fin = 721.876 uV 11:43:25 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -90.122 uVs 11:43:25 Integrated Voltage Measurement... 11:43:34 Voltage Offset Correction: v_off_ini = 735.013 uV, v_off_fin = 720.983 uV 11:43:34 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 125.520 uVs 11:43:37 I2X Average and RMS of Measurements VT = -109.657 +- 6.227 uVs I2X = 10.088 +- 0.573 uTm^2 11:43:39 Wire Position: X0 = 0.0000 mm, Y0 = -0.5000 mm 11:43:42 Integrated Voltage Measurement... 11:43:51 Voltage Offset Correction: v_off_ini = 714.793 uV, v_off_fin = 716.785 uV 11:43:51 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -346.917 uVs 11:43:51 Integrated Voltage Measurement... 11:44:00 Voltage Offset Correction: v_off_ini = 730.989 uV, v_off_fin = 720.937 uV 11:44:00 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 328.831 uVs 11:44:00 Integrated Voltage Measurement... 11:44:09 Voltage Offset Correction: v_off_ini = 726.406 uV, v_off_fin = 736.147 uV 11:44:09 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -336.668 uVs 11:44:09 Integrated Voltage Measurement... 11:44:18 Voltage Offset Correction: v_off_ini = 705.096 uV, v_off_fin = 721.505 uV 11:44:18 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 343.535 uVs 11:44:18 Integrated Voltage Measurement... 11:44:27 Voltage Offset Correction: v_off_ini = 707.671 uV, v_off_fin = 727.024 uV 11:44:27 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -334.882 uVs 11:44:27 Integrated Voltage Measurement... 11:44:36 Voltage Offset Correction: v_off_ini = 722.256 uV, v_off_fin = 741.452 uV 11:44:36 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 317.266 uVs 11:44:38 I2Y Average and RMS of Measurements VT = -334.683 +- 6.155 uVs I2Y = -30.791 +- 0.566 uTm^2 11:44:42 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:44:47 Integrated Voltage Measurement... 11:44:56 Voltage Offset Correction: v_off_ini = 730.060 uV, v_off_fin = 726.142 uV 11:44:56 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -82.701 uVs 11:44:56 Integrated Voltage Measurement... 11:45:05 Voltage Offset Correction: v_off_ini = 729.892 uV, v_off_fin = 738.854 uV 11:45:05 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 71.365 uVs 11:45:05 Integrated Voltage Measurement... 11:45:14 Voltage Offset Correction: v_off_ini = 734.428 uV, v_off_fin = 742.626 uV 11:45:14 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -53.788 uVs 11:45:14 Integrated Voltage Measurement... 11:45:23 Voltage Offset Correction: v_off_ini = 728.197 uV, v_off_fin = 731.417 uV 11:45:23 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 55.836 uVs 11:45:23 Integrated Voltage Measurement... 11:45:32 Voltage Offset Correction: v_off_ini = 729.231 uV, v_off_fin = 737.752 uV 11:45:32 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -72.715 uVs 11:45:32 Integrated Voltage Measurement... 11:45:41 Voltage Offset Correction: v_off_ini = 741.040 uV, v_off_fin = 754.330 uV 11:45:41 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 54.855 uVs 11:45:45 I1X Average and RMS of Measurements VT = -65.210 +- 9.128 uVs I1X = 13.042 +- 1.826 uTm 11:45:48 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:45:52 Integrated Voltage Measurement... 11:46:01 Voltage Offset Correction: v_off_ini = 724.823 uV, v_off_fin = 729.345 uV 11:46:01 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = -236.698 uVs 11:46:01 Integrated Voltage Measurement... 11:46:10 Voltage Offset Correction: v_off_ini = 749.400 uV, v_off_fin = 753.971 uV 11:46:10 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 243.710 uVs 11:46:10 Integrated Voltage Measurement... 11:46:19 Voltage Offset Correction: v_off_ini = 740.789 uV, v_off_fin = 718.766 uV 11:46:19 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -247.488 uVs 11:46:19 Integrated Voltage Measurement... 11:46:28 Voltage Offset Correction: v_off_ini = 749.393 uV, v_off_fin = 754.280 uV 11:46:28 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 253.814 uVs 11:46:28 Integrated Voltage Measurement... 11:46:37 Voltage Offset Correction: v_off_ini = 744.825 uV, v_off_fin = 729.738 uV 11:46:37 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -233.791 uVs 11:46:37 Integrated Voltage Measurement... 11:46:46 Voltage Offset Correction: v_off_ini = 716.086 uV, v_off_fin = 701.348 uV 11:46:46 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 246.331 uVs 11:46:50 I1Y Average and RMS of Measurements VT = -243.638 +- 4.959 uVs I1Y = -48.729 +- 0.992 uTm 11:46:53 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:46:55 Integrated Voltage Measurement... 11:47:04 Voltage Offset Correction: v_off_ini = 739.011 uV, v_off_fin = 724.454 uV 11:47:04 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -82.041 uVs 11:47:04 Integrated Voltage Measurement... 11:47:13 Voltage Offset Correction: v_off_ini = 732.434 uV, v_off_fin = 741.200 uV 11:47:13 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 100.590 uVs 11:47:13 Integrated Voltage Measurement... 11:47:22 Voltage Offset Correction: v_off_ini = 743.690 uV, v_off_fin = 730.086 uV 11:47:22 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -110.924 uVs 11:47:22 Integrated Voltage Measurement... 11:47:31 Voltage Offset Correction: v_off_ini = 697.263 uV, v_off_fin = 726.842 uV 11:47:31 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 116.909 uVs 11:47:31 Integrated Voltage Measurement... 11:47:40 Voltage Offset Correction: v_off_ini = 714.830 uV, v_off_fin = 721.228 uV 11:47:40 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -66.702 uVs 11:47:41 Integrated Voltage Measurement... 11:47:49 Voltage Offset Correction: v_off_ini = 740.624 uV, v_off_fin = 749.181 uV 11:47:50 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 85.614 uVs 11:47:52 I2X Average and RMS of Measurements VT = -93.797 +- 15.514 uVs I2X = 8.629 +- 1.427 uTm^2 11:47:54 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 11:47:57 Integrated Voltage Measurement... 11:48:06 Voltage Offset Correction: v_off_ini = 721.779 uV, v_off_fin = 726.500 uV 11:48:06 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -340.827 uVs 11:48:06 Integrated Voltage Measurement... 11:48:15 Voltage Offset Correction: v_off_ini = 714.467 uV, v_off_fin = 722.025 uV 11:48:15 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 355.748 uVs 11:48:15 Integrated Voltage Measurement... 11:48:24 Voltage Offset Correction: v_off_ini = 732.083 uV, v_off_fin = 717.712 uV 11:48:24 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -335.659 uVs 11:48:24 Integrated Voltage Measurement... 11:48:33 Voltage Offset Correction: v_off_ini = 733.339 uV, v_off_fin = 727.960 uV 11:48:33 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 319.657 uVs 11:48:33 Integrated Voltage Measurement... 11:48:42 Voltage Offset Correction: v_off_ini = 744.524 uV, v_off_fin = 731.759 uV 11:48:42 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -322.430 uVs 11:48:42 Integrated Voltage Measurement... 11:48:51 Voltage Offset Correction: v_off_ini = 719.390 uV, v_off_fin = 724.477 uV 11:48:51 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 338.856 uVs 11:48:54 I2Y Average and RMS of Measurements VT = -335.529 +- 9.103 uVs I2Y = -30.869 +- 0.838 uTm^2 11:48:58 Wire Position: X0 = 0.0000 mm, Y0 = 0.5000 mm 11:49:02 Integrated Voltage Measurement... 11:49:11 Voltage Offset Correction: v_off_ini = 721.028 uV, v_off_fin = 715.927 uV 11:49:11 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -56.634 uVs 11:49:11 Integrated Voltage Measurement... 11:49:20 Voltage Offset Correction: v_off_ini = 730.082 uV, v_off_fin = 721.719 uV 11:49:20 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 90.257 uVs 11:49:20 Integrated Voltage Measurement... 11:49:29 Voltage Offset Correction: v_off_ini = 726.664 uV, v_off_fin = 732.779 uV 11:49:29 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -66.007 uVs 11:49:29 Integrated Voltage Measurement... 11:49:38 Voltage Offset Correction: v_off_ini = 727.043 uV, v_off_fin = 719.020 uV 11:49:38 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 61.360 uVs 11:49:38 Integrated Voltage Measurement... 11:49:47 Voltage Offset Correction: v_off_ini = 720.636 uV, v_off_fin = 736.233 uV 11:49:47 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -64.027 uVs 11:49:47 Integrated Voltage Measurement... 11:49:56 Voltage Offset Correction: v_off_ini = 728.239 uV, v_off_fin = 720.605 uV 11:49:56 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 97.396 uVs 11:50:00 I1X Average and RMS of Measurements VT = -72.613 +- 6.977 uVs I1X = 14.523 +- 1.395 uTm 11:50:03 Wire Position: X0 = 0.0000 mm, Y0 = 0.5000 mm 11:50:07 Integrated Voltage Measurement... 11:50:16 Voltage Offset Correction: v_off_ini = 716.487 uV, v_off_fin = 722.062 uV 11:50:16 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -242.452 uVs 11:50:16 Integrated Voltage Measurement... 11:50:25 Voltage Offset Correction: v_off_ini = 721.363 uV, v_off_fin = 740.342 uV 11:50:25 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 238.121 uVs 11:50:25 Integrated Voltage Measurement... 11:50:34 Voltage Offset Correction: v_off_ini = 720.616 uV, v_off_fin = 724.210 uV 11:50:34 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -250.086 uVs 11:50:34 Integrated Voltage Measurement... 11:50:43 Voltage Offset Correction: v_off_ini = 740.793 uV, v_off_fin = 749.904 uV 11:50:43 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 243.022 uVs 11:50:43 Integrated Voltage Measurement... 11:50:52 Voltage Offset Correction: v_off_ini = 728.874 uV, v_off_fin = 716.234 uV 11:50:52 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5002 mm dXY = 0.5001 mm VT = -251.025 uVs 11:50:52 Integrated Voltage Measurement... 11:51:01 Voltage Offset Correction: v_off_ini = 721.863 uV, v_off_fin = 727.677 uV 11:51:01 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 242.254 uVs 11:51:05 I1Y Average and RMS of Measurements VT = -244.493 +- 2.975 uVs I1Y = -48.895 +- 0.593 uTm 11:51:08 Wire Position: X0 = 0.0000 mm, Y0 = 0.5000 mm 11:51:10 Integrated Voltage Measurement... 11:51:19 Voltage Offset Correction: v_off_ini = 723.455 uV, v_off_fin = 741.758 uV 11:51:20 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -84.343 uVs 11:51:20 Integrated Voltage Measurement... 11:51:29 Voltage Offset Correction: v_off_ini = 750.764 uV, v_off_fin = 745.441 uV 11:51:29 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 96.993 uVs 11:51:29 Integrated Voltage Measurement... 11:51:38 Voltage Offset Correction: v_off_ini = 724.760 uV, v_off_fin = 711.705 uV 11:51:38 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -77.536 uVs 11:51:38 Integrated Voltage Measurement... 11:51:47 Voltage Offset Correction: v_off_ini = 730.141 uV, v_off_fin = 735.420 uV 11:51:47 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 105.274 uVs 11:51:47 Integrated Voltage Measurement... 11:51:56 Voltage Offset Correction: v_off_ini = 722.001 uV, v_off_fin = 704.460 uV 11:51:56 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -50.868 uVs 11:51:56 Integrated Voltage Measurement... 11:52:05 Voltage Offset Correction: v_off_ini = 708.067 uV, v_off_fin = 714.153 uV 11:52:05 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 88.093 uVs 11:52:07 I2X Average and RMS of Measurements VT = -83.851 +- 10.166 uVs I2X = 7.714 +- 0.935 uTm^2 11:52:10 Wire Position: X0 = 0.0000 mm, Y0 = 0.5000 mm 11:52:12 Integrated Voltage Measurement... 11:52:21 Voltage Offset Correction: v_off_ini = 719.896 uV, v_off_fin = 725.467 uV 11:52:21 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -328.942 uVs 11:52:21 Integrated Voltage Measurement... 11:52:30 Voltage Offset Correction: v_off_ini = 737.159 uV, v_off_fin = 734.054 uV 11:52:30 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 351.706 uVs 11:52:30 Integrated Voltage Measurement... 11:52:39 Voltage Offset Correction: v_off_ini = 742.734 uV, v_off_fin = 735.219 uV 11:52:39 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -321.295 uVs 11:52:39 Integrated Voltage Measurement... 11:52:48 Voltage Offset Correction: v_off_ini = 733.632 uV, v_off_fin = 754.607 uV 11:52:48 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 340.766 uVs 11:52:48 Integrated Voltage Measurement... 11:52:57 Voltage Offset Correction: v_off_ini = 743.515 uV, v_off_fin = 737.391 uV 11:52:57 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -343.968 uVs 11:52:57 Integrated Voltage Measurement... 11:53:06 Voltage Offset Correction: v_off_ini = 720.540 uV, v_off_fin = 698.905 uV 11:53:06 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 327.478 uVs 11:53:09 I2Y Average and RMS of Measurements VT = -335.692 +- 3.794 uVs I2Y = -30.884 +- 0.349 uTm^2 11:53:13 Wire Position: X0 = 0.0000 mm, Y0 = 1.0000 mm 11:53:17 Integrated Voltage Measurement... 11:53:26 Voltage Offset Correction: v_off_ini = 713.100 uV, v_off_fin = 717.512 uV 11:53:26 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -46.653 uVs 11:53:26 Integrated Voltage Measurement... 11:53:35 Voltage Offset Correction: v_off_ini = 728.394 uV, v_off_fin = 708.810 uV 11:53:35 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 77.158 uVs 11:53:35 Integrated Voltage Measurement... 11:53:44 Voltage Offset Correction: v_off_ini = 710.654 uV, v_off_fin = 714.600 uV 11:53:44 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -52.321 uVs 11:53:44 Integrated Voltage Measurement... 11:53:53 Voltage Offset Correction: v_off_ini = 727.148 uV, v_off_fin = 728.736 uV 11:53:53 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 71.585 uVs 11:53:53 Integrated Voltage Measurement... 11:54:02 Voltage Offset Correction: v_off_ini = 720.923 uV, v_off_fin = 727.129 uV 11:54:02 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -65.993 uVs 11:54:02 Integrated Voltage Measurement... 11:54:11 Voltage Offset Correction: v_off_ini = 744.526 uV, v_off_fin = 712.418 uV 11:54:11 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 63.510 uVs 11:54:16 I1X Average and RMS of Measurements VT = -62.870 +- 1.331 uVs I1X = 12.574 +- 0.266 uTm 11:54:18 Wire Position: X0 = 0.0000 mm, Y0 = 1.0000 mm 11:54:22 Integrated Voltage Measurement... 11:54:31 Voltage Offset Correction: v_off_ini = 714.724 uV, v_off_fin = 719.152 uV 11:54:31 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = -223.334 uVs 11:54:31 Integrated Voltage Measurement... 11:54:40 Voltage Offset Correction: v_off_ini = 743.690 uV, v_off_fin = 762.350 uV 11:54:40 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 231.494 uVs 11:54:40 Integrated Voltage Measurement... 11:54:49 Voltage Offset Correction: v_off_ini = 720.499 uV, v_off_fin = 681.442 uV 11:54:49 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -240.300 uVs 11:54:49 Integrated Voltage Measurement... 11:54:58 Voltage Offset Correction: v_off_ini = 697.466 uV, v_off_fin = 723.439 uV 11:54:58 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 240.706 uVs 11:54:58 Integrated Voltage Measurement... 11:55:07 Voltage Offset Correction: v_off_ini = 711.313 uV, v_off_fin = 713.291 uV 11:55:07 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -240.164 uVs 11:55:07 Integrated Voltage Measurement... 11:55:16 Voltage Offset Correction: v_off_ini = 723.549 uV, v_off_fin = 753.441 uV 11:55:16 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 240.406 uVs 11:55:20 I1Y Average and RMS of Measurements VT = -236.067 +- 6.119 uVs I1Y = -47.210 +- 1.222 uTm 11:55:23 Wire Position: X0 = 0.0000 mm, Y0 = 1.0000 mm 11:55:26 Integrated Voltage Measurement... 11:55:35 Voltage Offset Correction: v_off_ini = 742.340 uV, v_off_fin = 719.416 uV 11:55:35 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -70.057 uVs 11:55:35 Integrated Voltage Measurement... 11:55:44 Voltage Offset Correction: v_off_ini = 740.946 uV, v_off_fin = 738.492 uV 11:55:44 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 107.043 uVs 11:55:44 Integrated Voltage Measurement... 11:55:53 Voltage Offset Correction: v_off_ini = 730.292 uV, v_off_fin = 718.316 uV 11:55:53 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -70.112 uVs 11:55:53 Integrated Voltage Measurement... 11:56:02 Voltage Offset Correction: v_off_ini = 726.787 uV, v_off_fin = 738.219 uV 11:56:02 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 90.672 uVs 11:56:02 Integrated Voltage Measurement... 11:56:11 Voltage Offset Correction: v_off_ini = 724.558 uV, v_off_fin = 714.107 uV 11:56:11 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -102.917 uVs 11:56:11 Integrated Voltage Measurement... 11:56:20 Voltage Offset Correction: v_off_ini = 730.639 uV, v_off_fin = 735.796 uV 11:56:20 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 94.925 uVs 11:56:22 I2X Average and RMS of Measurements VT = -89.288 +- 7.582 uVs I2X = 8.215 +- 0.698 uTm^2 11:56:25 Wire Position: X0 = 0.0000 mm, Y0 = 1.0000 mm 11:56:27 Integrated Voltage Measurement... 11:56:36 Voltage Offset Correction: v_off_ini = 721.756 uV, v_off_fin = 721.366 uV 11:56:36 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -335.044 uVs 11:56:36 Integrated Voltage Measurement... 11:56:45 Voltage Offset Correction: v_off_ini = 722.760 uV, v_off_fin = 736.463 uV 11:56:45 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 329.969 uVs 11:56:45 Integrated Voltage Measurement... 11:56:54 Voltage Offset Correction: v_off_ini = 713.648 uV, v_off_fin = 714.247 uV 11:56:54 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -328.826 uVs 11:56:54 Integrated Voltage Measurement... 11:57:03 Voltage Offset Correction: v_off_ini = 724.477 uV, v_off_fin = 711.959 uV 11:57:03 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = 347.477 uVs 11:57:04 Integrated Voltage Measurement... 11:57:13 Voltage Offset Correction: v_off_ini = 725.784 uV, v_off_fin = 713.712 uV 11:57:13 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -333.286 uVs 11:57:13 Integrated Voltage Measurement... 11:57:22 Voltage Offset Correction: v_off_ini = 727.189 uV, v_off_fin = 716.279 uV 11:57:22 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 334.748 uVs 11:57:24 I2Y Average and RMS of Measurements VT = -334.892 +- 2.386 uVs I2Y = -30.811 +- 0.220 uTm^2 11:57:39 Wire Position: X0 = -2.0000 mm, Y0 = 0.0000 mm 11:57:43 Integrated Voltage Measurement... 11:57:52 Voltage Offset Correction: v_off_ini = 735.719 uV, v_off_fin = 735.743 uV 11:57:52 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -55.261 uVs 11:57:52 Integrated Voltage Measurement... 11:58:01 Voltage Offset Correction: v_off_ini = 742.725 uV, v_off_fin = 724.319 uV 11:58:01 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 82.310 uVs 11:58:01 Integrated Voltage Measurement... 11:58:10 Voltage Offset Correction: v_off_ini = 731.551 uV, v_off_fin = 721.798 uV 11:58:10 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -46.508 uVs 11:58:10 Integrated Voltage Measurement... 11:58:19 Voltage Offset Correction: v_off_ini = 734.145 uV, v_off_fin = 732.683 uV 11:58:19 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 55.680 uVs 11:58:19 Integrated Voltage Measurement... 11:58:28 Voltage Offset Correction: v_off_ini = 754.651 uV, v_off_fin = 738.633 uV 11:58:28 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -55.566 uVs 11:58:28 Integrated Voltage Measurement... 11:58:37 Voltage Offset Correction: v_off_ini = 740.679 uV, v_off_fin = 748.993 uV 11:58:37 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 67.644 uVs 11:58:41 I1X Average and RMS of Measurements VT = -60.495 +- 7.265 uVs I1X = 12.099 +- 1.453 uTm 11:58:44 Wire Position: X0 = -2.0000 mm, Y0 = 0.0000 mm 11:58:48 Integrated Voltage Measurement... 11:58:57 Voltage Offset Correction: v_off_ini = 728.546 uV, v_off_fin = 706.557 uV 11:58:57 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -243.745 uVs 11:58:57 Integrated Voltage Measurement... 11:59:06 Voltage Offset Correction: v_off_ini = 723.220 uV, v_off_fin = 746.232 uV 11:59:06 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 268.979 uVs 11:59:06 Integrated Voltage Measurement... 11:59:15 Voltage Offset Correction: v_off_ini = 741.723 uV, v_off_fin = 731.865 uV 11:59:15 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -241.096 uVs 11:59:15 Integrated Voltage Measurement... 11:59:24 Voltage Offset Correction: v_off_ini = 739.029 uV, v_off_fin = 747.925 uV 11:59:24 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 230.089 uVs 11:59:24 Integrated Voltage Measurement... 11:59:33 Voltage Offset Correction: v_off_ini = 723.952 uV, v_off_fin = 715.253 uV 11:59:33 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -236.206 uVs 11:59:33 Integrated Voltage Measurement... 11:59:42 Voltage Offset Correction: v_off_ini = 712.167 uV, v_off_fin = 726.228 uV 11:59:42 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 242.808 uVs 11:59:46 I1Y Average and RMS of Measurements VT = -243.820 +- 9.011 uVs I1Y = -48.764 +- 1.802 uTm 11:59:51 Wire Position: X0 = -2.0000 mm, Y0 = 0.0000 mm 11:59:54 Integrated Voltage Measurement... 12:00:03 Voltage Offset Correction: v_off_ini = 737.366 uV, v_off_fin = 725.707 uV 12:00:03 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -81.912 uVs 12:00:03 Integrated Voltage Measurement... 12:00:12 Voltage Offset Correction: v_off_ini = 741.194 uV, v_off_fin = 743.296 uV 12:00:12 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 95.764 uVs 12:00:12 Integrated Voltage Measurement... 12:00:21 Voltage Offset Correction: v_off_ini = 720.724 uV, v_off_fin = 727.770 uV 12:00:21 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -98.126 uVs 12:00:21 Integrated Voltage Measurement... 12:00:30 Voltage Offset Correction: v_off_ini = 746.802 uV, v_off_fin = 718.813 uV 12:00:30 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 70.268 uVs 12:00:30 Integrated Voltage Measurement... 12:00:39 Voltage Offset Correction: v_off_ini = 705.609 uV, v_off_fin = 707.827 uV 12:00:39 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -80.722 uVs 12:00:39 Integrated Voltage Measurement... 12:00:48 Voltage Offset Correction: v_off_ini = 718.351 uV, v_off_fin = 725.584 uV 12:00:48 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 75.909 uVs 12:00:51 I2X Average and RMS of Measurements VT = -83.784 +- 4.306 uVs I2X = 7.708 +- 0.396 uTm^2 12:00:56 Wire Position: X0 = -2.0000 mm, Y0 = 0.0000 mm 12:00:58 Integrated Voltage Measurement... 12:01:07 Voltage Offset Correction: v_off_ini = 720.342 uV, v_off_fin = 722.858 uV 12:01:07 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -333.016 uVs 12:01:07 Integrated Voltage Measurement... 12:01:16 Voltage Offset Correction: v_off_ini = 739.092 uV, v_off_fin = 728.029 uV 12:01:16 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 311.217 uVs 12:01:16 Integrated Voltage Measurement... 12:01:25 Voltage Offset Correction: v_off_ini = 730.405 uV, v_off_fin = 723.841 uV 12:01:26 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -327.450 uVs 12:01:26 Integrated Voltage Measurement... 12:01:35 Voltage Offset Correction: v_off_ini = 721.447 uV, v_off_fin = 732.413 uV 12:01:35 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 334.406 uVs 12:01:35 Integrated Voltage Measurement... 12:01:44 Voltage Offset Correction: v_off_ini = 728.633 uV, v_off_fin = 718.299 uV 12:01:44 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -348.737 uVs 12:01:44 Integrated Voltage Measurement... 12:01:53 Voltage Offset Correction: v_off_ini = 732.332 uV, v_off_fin = 726.127 uV 12:01:53 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 331.181 uVs 12:01:55 I2Y Average and RMS of Measurements VT = -331.001 +- 7.284 uVs I2Y = -30.452 +- 0.670 uTm^2 12:02:00 Wire Position: X0 = -1.0000 mm, Y0 = 0.0000 mm 12:02:04 Integrated Voltage Measurement... 12:02:13 Voltage Offset Correction: v_off_ini = 731.969 uV, v_off_fin = 741.583 uV 12:02:13 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.4998 mm dXY = -0.4999 mm VT = -69.735 uVs 12:02:13 Integrated Voltage Measurement... 12:02:22 Voltage Offset Correction: v_off_ini = 725.832 uV, v_off_fin = 730.465 uV 12:02:22 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 81.970 uVs 12:02:22 Integrated Voltage Measurement... 12:02:31 Voltage Offset Correction: v_off_ini = 740.323 uV, v_off_fin = 753.042 uV 12:02:31 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -81.025 uVs 12:02:31 Integrated Voltage Measurement... 12:02:40 Voltage Offset Correction: v_off_ini = 728.242 uV, v_off_fin = 736.074 uV 12:02:40 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 49.346 uVs 12:02:40 Integrated Voltage Measurement... 12:02:49 Voltage Offset Correction: v_off_ini = 739.806 uV, v_off_fin = 735.397 uV 12:02:49 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -80.868 uVs 12:02:49 Integrated Voltage Measurement... 12:02:58 Voltage Offset Correction: v_off_ini = 715.958 uV, v_off_fin = 726.360 uV 12:02:58 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 59.418 uVs 12:03:02 I1X Average and RMS of Measurements VT = -70.394 +- 4.358 uVs I1X = 14.079 +- 0.873 uTm 12:03:05 Wire Position: X0 = -1.0000 mm, Y0 = 0.0000 mm 12:03:09 Integrated Voltage Measurement... 12:03:18 Voltage Offset Correction: v_off_ini = 712.549 uV, v_off_fin = 693.073 uV 12:03:18 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -248.158 uVs 12:03:18 Integrated Voltage Measurement... 12:03:27 Voltage Offset Correction: v_off_ini = 709.897 uV, v_off_fin = 734.786 uV 12:03:27 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 242.005 uVs 12:03:27 Integrated Voltage Measurement... 12:03:36 Voltage Offset Correction: v_off_ini = 738.456 uV, v_off_fin = 721.534 uV 12:03:36 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -255.376 uVs 12:03:36 Integrated Voltage Measurement... 12:03:45 Voltage Offset Correction: v_off_ini = 714.536 uV, v_off_fin = 704.657 uV 12:03:45 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 239.110 uVs 12:03:45 Integrated Voltage Measurement... 12:03:54 Voltage Offset Correction: v_off_ini = 722.800 uV, v_off_fin = 706.645 uV 12:03:54 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -242.683 uVs 12:03:54 Integrated Voltage Measurement... 12:04:03 Voltage Offset Correction: v_off_ini = 720.241 uV, v_off_fin = 737.087 uV 12:04:03 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 256.192 uVs 12:04:07 I1Y Average and RMS of Measurements VT = -247.254 +- 1.778 uVs I1Y = -49.451 +- 0.356 uTm 12:04:12 Wire Position: X0 = -1.0000 mm, Y0 = 0.0000 mm 12:04:15 Integrated Voltage Measurement... 12:04:24 Voltage Offset Correction: v_off_ini = 735.025 uV, v_off_fin = 738.709 uV 12:04:24 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -79.493 uVs 12:04:24 Integrated Voltage Measurement... 12:04:33 Voltage Offset Correction: v_off_ini = 722.832 uV, v_off_fin = 731.949 uV 12:04:33 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 74.088 uVs 12:04:33 Integrated Voltage Measurement... 12:04:42 Voltage Offset Correction: v_off_ini = 732.736 uV, v_off_fin = 709.763 uV 12:04:42 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -81.761 uVs 12:04:42 Integrated Voltage Measurement... 12:04:51 Voltage Offset Correction: v_off_ini = 731.770 uV, v_off_fin = 722.808 uV 12:04:51 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 71.629 uVs 12:04:51 Integrated Voltage Measurement... 12:05:00 Voltage Offset Correction: v_off_ini = 731.687 uV, v_off_fin = 724.964 uV 12:05:00 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -110.944 uVs 12:05:00 Integrated Voltage Measurement... 12:05:09 Voltage Offset Correction: v_off_ini = 723.692 uV, v_off_fin = 733.365 uV 12:05:09 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 91.609 uVs 12:05:12 I2X Average and RMS of Measurements VT = -84.921 +- 11.565 uVs I2X = 7.813 +- 1.064 uTm^2 12:05:17 Wire Position: X0 = -1.0000 mm, Y0 = 0.0000 mm 12:05:19 Integrated Voltage Measurement... 12:05:28 Voltage Offset Correction: v_off_ini = 723.280 uV, v_off_fin = 723.471 uV 12:05:28 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -327.176 uVs 12:05:28 Integrated Voltage Measurement... 12:05:37 Voltage Offset Correction: v_off_ini = 713.727 uV, v_off_fin = 726.920 uV 12:05:37 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 324.246 uVs 12:05:37 Integrated Voltage Measurement... 12:05:46 Voltage Offset Correction: v_off_ini = 718.888 uV, v_off_fin = 694.805 uV 12:05:46 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -328.218 uVs 12:05:47 Integrated Voltage Measurement... 12:05:56 Voltage Offset Correction: v_off_ini = 707.508 uV, v_off_fin = 708.664 uV 12:05:56 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 311.389 uVs 12:05:56 Integrated Voltage Measurement... 12:06:05 Voltage Offset Correction: v_off_ini = 693.089 uV, v_off_fin = 691.457 uV 12:06:05 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -337.579 uVs 12:06:05 Integrated Voltage Measurement... 12:06:14 Voltage Offset Correction: v_off_ini = 720.484 uV, v_off_fin = 727.252 uV 12:06:14 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 314.281 uVs 12:06:16 I2Y Average and RMS of Measurements VT = -323.815 +- 2.838 uVs I2Y = -29.791 +- 0.261 uTm^2 12:06:21 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 12:06:25 Integrated Voltage Measurement... 12:06:34 Voltage Offset Correction: v_off_ini = 709.466 uV, v_off_fin = 720.361 uV 12:06:34 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -44.856 uVs 12:06:34 Integrated Voltage Measurement... 12:06:43 Voltage Offset Correction: v_off_ini = 718.863 uV, v_off_fin = 719.923 uV 12:06:43 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 69.021 uVs 12:06:43 Integrated Voltage Measurement... 12:06:52 Voltage Offset Correction: v_off_ini = 729.035 uV, v_off_fin = 724.144 uV 12:06:52 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -88.942 uVs 12:06:52 Integrated Voltage Measurement... 12:07:01 Voltage Offset Correction: v_off_ini = 719.237 uV, v_off_fin = 719.602 uV 12:07:01 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 82.122 uVs 12:07:01 Integrated Voltage Measurement... 12:07:10 Voltage Offset Correction: v_off_ini = 736.237 uV, v_off_fin = 748.152 uV 12:07:10 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -81.645 uVs 12:07:10 Integrated Voltage Measurement... 12:07:19 Voltage Offset Correction: v_off_ini = 737.776 uV, v_off_fin = 729.556 uV 12:07:19 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 68.900 uVs 12:07:23 I1X Average and RMS of Measurements VT = -72.581 +- 11.827 uVs I1X = 14.516 +- 2.365 uTm 12:07:26 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 12:07:30 Integrated Voltage Measurement... 12:07:39 Voltage Offset Correction: v_off_ini = 734.464 uV, v_off_fin = 703.187 uV 12:07:39 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = -241.361 uVs 12:07:39 Integrated Voltage Measurement... 12:07:48 Voltage Offset Correction: v_off_ini = 717.899 uV, v_off_fin = 728.016 uV 12:07:48 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 245.129 uVs 12:07:48 Integrated Voltage Measurement... 12:07:57 Voltage Offset Correction: v_off_ini = 733.427 uV, v_off_fin = 732.592 uV 12:07:57 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -248.017 uVs 12:07:57 Integrated Voltage Measurement... 12:08:06 Voltage Offset Correction: v_off_ini = 720.572 uV, v_off_fin = 738.772 uV 12:08:06 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 248.635 uVs 12:08:06 Integrated Voltage Measurement... 12:08:15 Voltage Offset Correction: v_off_ini = 732.583 uV, v_off_fin = 721.957 uV 12:08:15 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -242.358 uVs 12:08:15 Integrated Voltage Measurement... 12:08:24 Voltage Offset Correction: v_off_ini = 742.574 uV, v_off_fin = 750.013 uV 12:08:24 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 230.750 uVs 12:08:28 I1Y Average and RMS of Measurements VT = -242.708 +- 4.821 uVs I1Y = -48.542 +- 0.964 uTm 12:08:33 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 12:08:36 Integrated Voltage Measurement... 12:08:45 Voltage Offset Correction: v_off_ini = 716.417 uV, v_off_fin = 712.646 uV 12:08:45 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -106.472 uVs 12:08:45 Integrated Voltage Measurement... 12:08:54 Voltage Offset Correction: v_off_ini = 733.285 uV, v_off_fin = 750.688 uV 12:08:54 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 83.395 uVs 12:08:54 Integrated Voltage Measurement... 12:09:03 Voltage Offset Correction: v_off_ini = 741.430 uV, v_off_fin = 711.985 uV 12:09:03 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -90.679 uVs 12:09:03 Integrated Voltage Measurement... 12:09:12 Voltage Offset Correction: v_off_ini = 721.888 uV, v_off_fin = 734.422 uV 12:09:12 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 111.792 uVs 12:09:12 Integrated Voltage Measurement... 12:09:21 Voltage Offset Correction: v_off_ini = 730.641 uV, v_off_fin = 733.117 uV 12:09:21 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -104.335 uVs 12:09:21 Integrated Voltage Measurement... 12:09:30 Voltage Offset Correction: v_off_ini = 718.913 uV, v_off_fin = 731.832 uV 12:09:30 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 77.664 uVs 12:09:33 I2X Average and RMS of Measurements VT = -95.723 +- 4.216 uVs I2X = 8.807 +- 0.388 uTm^2 12:09:38 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 12:09:40 Integrated Voltage Measurement... 12:09:49 Voltage Offset Correction: v_off_ini = 714.707 uV, v_off_fin = 711.770 uV 12:09:49 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -357.714 uVs 12:09:49 Integrated Voltage Measurement... 12:09:58 Voltage Offset Correction: v_off_ini = 706.695 uV, v_off_fin = 722.825 uV 12:09:58 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 305.026 uVs 12:09:59 Integrated Voltage Measurement... 12:10:07 Voltage Offset Correction: v_off_ini = 702.281 uV, v_off_fin = 694.496 uV 12:10:08 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -346.352 uVs 12:10:08 Integrated Voltage Measurement... 12:10:17 Voltage Offset Correction: v_off_ini = 731.910 uV, v_off_fin = 704.979 uV 12:10:17 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 339.703 uVs 12:10:17 Integrated Voltage Measurement... 12:10:26 Voltage Offset Correction: v_off_ini = 692.704 uV, v_off_fin = 716.584 uV 12:10:26 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -316.110 uVs 12:10:26 Integrated Voltage Measurement... 12:10:35 Voltage Offset Correction: v_off_ini = 709.945 uV, v_off_fin = 707.176 uV 12:10:35 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 341.678 uVs 12:10:37 I2Y Average and RMS of Measurements VT = -334.431 +- 6.162 uVs I2Y = -30.768 +- 0.567 uTm^2 12:10:42 Wire Position: X0 = 1.0000 mm, Y0 = 0.0000 mm 12:10:46 Integrated Voltage Measurement... 12:10:55 Voltage Offset Correction: v_off_ini = 735.936 uV, v_off_fin = 710.294 uV 12:10:55 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -61.002 uVs 12:10:55 Integrated Voltage Measurement... 12:11:04 Voltage Offset Correction: v_off_ini = 724.201 uV, v_off_fin = 723.680 uV 12:11:04 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 77.957 uVs 12:11:04 Integrated Voltage Measurement... 12:11:13 Voltage Offset Correction: v_off_ini = 705.132 uV, v_off_fin = 725.516 uV 12:11:13 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -65.383 uVs 12:11:13 Integrated Voltage Measurement... 12:11:22 Voltage Offset Correction: v_off_ini = 729.500 uV, v_off_fin = 718.429 uV 12:11:22 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 110.229 uVs 12:11:22 Integrated Voltage Measurement... 12:11:31 Voltage Offset Correction: v_off_ini = 737.687 uV, v_off_fin = 736.837 uV 12:11:31 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -53.526 uVs 12:11:31 Integrated Voltage Measurement... 12:11:40 Voltage Offset Correction: v_off_ini = 721.795 uV, v_off_fin = 726.040 uV 12:11:40 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 50.661 uVs 12:11:44 I1X Average and RMS of Measurements VT = -69.793 +- 14.581 uVs I1X = 13.959 +- 2.916 uTm 12:11:47 Wire Position: X0 = 1.0000 mm, Y0 = 0.0000 mm 12:11:51 Integrated Voltage Measurement... 12:12:00 Voltage Offset Correction: v_off_ini = 719.763 uV, v_off_fin = 709.112 uV 12:12:00 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -243.460 uVs 12:12:00 Integrated Voltage Measurement... 12:12:09 Voltage Offset Correction: v_off_ini = 724.774 uV, v_off_fin = 754.326 uV 12:12:09 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 253.393 uVs 12:12:09 Integrated Voltage Measurement... 12:12:18 Voltage Offset Correction: v_off_ini = 723.415 uV, v_off_fin = 713.441 uV 12:12:18 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -260.066 uVs 12:12:18 Integrated Voltage Measurement... 12:12:27 Voltage Offset Correction: v_off_ini = 733.722 uV, v_off_fin = 745.220 uV 12:12:27 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 261.831 uVs 12:12:27 Integrated Voltage Measurement... 12:12:36 Voltage Offset Correction: v_off_ini = 734.016 uV, v_off_fin = 725.632 uV 12:12:36 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -251.445 uVs 12:12:36 Integrated Voltage Measurement... 12:12:45 Voltage Offset Correction: v_off_ini = 731.594 uV, v_off_fin = 708.820 uV 12:12:45 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 266.260 uVs 12:12:49 I1Y Average and RMS of Measurements VT = -256.076 +- 5.476 uVs I1Y = -51.215 +- 1.095 uTm 12:12:54 Wire Position: X0 = 1.0000 mm, Y0 = 0.0000 mm 12:12:57 Integrated Voltage Measurement... 12:13:06 Voltage Offset Correction: v_off_ini = 725.322 uV, v_off_fin = 722.975 uV 12:13:06 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -82.485 uVs 12:13:06 Integrated Voltage Measurement... 12:13:15 Voltage Offset Correction: v_off_ini = 734.943 uV, v_off_fin = 744.902 uV 12:13:15 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 100.700 uVs 12:13:15 Integrated Voltage Measurement... 12:13:24 Voltage Offset Correction: v_off_ini = 733.611 uV, v_off_fin = 729.117 uV 12:13:24 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -109.307 uVs 12:13:24 Integrated Voltage Measurement... 12:13:33 Voltage Offset Correction: v_off_ini = 732.439 uV, v_off_fin = 746.541 uV 12:13:33 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 89.722 uVs 12:13:33 Integrated Voltage Measurement... 12:13:42 Voltage Offset Correction: v_off_ini = 725.660 uV, v_off_fin = 733.235 uV 12:13:42 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -64.863 uVs 12:13:42 Integrated Voltage Measurement... 12:13:51 Voltage Offset Correction: v_off_ini = 734.175 uV, v_off_fin = 751.955 uV 12:13:51 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 95.499 uVs 12:13:54 I2X Average and RMS of Measurements VT = -90.429 +- 7.936 uVs I2X = 8.319 +- 0.730 uTm^2 12:13:59 Wire Position: X0 = 1.0000 mm, Y0 = 0.0000 mm 12:14:01 Integrated Voltage Measurement... 12:14:10 Voltage Offset Correction: v_off_ini = 754.151 uV, v_off_fin = 757.860 uV 12:14:10 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -323.248 uVs 12:14:10 Integrated Voltage Measurement... 12:14:19 Voltage Offset Correction: v_off_ini = 747.157 uV, v_off_fin = 737.192 uV 12:14:20 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 337.098 uVs 12:14:20 Integrated Voltage Measurement... 12:14:29 Voltage Offset Correction: v_off_ini = 740.331 uV, v_off_fin = 752.746 uV 12:14:29 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -345.314 uVs 12:14:29 Integrated Voltage Measurement... 12:14:38 Voltage Offset Correction: v_off_ini = 743.376 uV, v_off_fin = 737.589 uV 12:14:38 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = 348.709 uVs 12:14:38 Integrated Voltage Measurement... 12:14:47 Voltage Offset Correction: v_off_ini = 736.145 uV, v_off_fin = 747.385 uV 12:14:47 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -349.336 uVs 12:14:47 Integrated Voltage Measurement... 12:14:56 Voltage Offset Correction: v_off_ini = 769.694 uV, v_off_fin = 756.747 uV 12:14:56 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 322.511 uVs 12:14:58 I2Y Average and RMS of Measurements VT = -337.703 +- 6.988 uVs I2Y = -31.069 +- 0.644 uTm^2 12:15:03 Wire Position: X0 = 2.0000 mm, Y0 = 0.0000 mm 12:15:07 Integrated Voltage Measurement... 12:15:16 Voltage Offset Correction: v_off_ini = 750.586 uV, v_off_fin = 752.748 uV 12:15:16 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -88.076 uVs 12:15:16 Integrated Voltage Measurement... 12:15:25 Voltage Offset Correction: v_off_ini = 733.862 uV, v_off_fin = 749.893 uV 12:15:25 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 68.664 uVs 12:15:25 Integrated Voltage Measurement... 12:15:34 Voltage Offset Correction: v_off_ini = 726.471 uV, v_off_fin = 747.322 uV 12:15:34 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -72.964 uVs 12:15:34 Integrated Voltage Measurement... 12:15:43 Voltage Offset Correction: v_off_ini = 731.904 uV, v_off_fin = 751.169 uV 12:15:43 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 81.053 uVs 12:15:43 Integrated Voltage Measurement... 12:15:52 Voltage Offset Correction: v_off_ini = 762.443 uV, v_off_fin = 773.831 uV 12:15:52 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -84.496 uVs 12:15:52 Integrated Voltage Measurement... 12:16:01 Voltage Offset Correction: v_off_ini = 752.041 uV, v_off_fin = 751.366 uV 12:16:01 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 76.977 uVs 12:16:05 I1X Average and RMS of Measurements VT = -78.705 +- 1.540 uVs I1X = 15.741 +- 0.308 uTm 12:16:08 Wire Position: X0 = 2.0000 mm, Y0 = 0.0000 mm 12:16:12 Integrated Voltage Measurement... 12:16:21 Voltage Offset Correction: v_off_ini = 729.399 uV, v_off_fin = 714.356 uV 12:16:21 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -246.899 uVs 12:16:21 Integrated Voltage Measurement... 12:16:30 Voltage Offset Correction: v_off_ini = 731.258 uV, v_off_fin = 743.640 uV 12:16:30 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 275.459 uVs 12:16:30 Integrated Voltage Measurement... 12:16:39 Voltage Offset Correction: v_off_ini = 723.362 uV, v_off_fin = 694.936 uV 12:16:39 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -267.516 uVs 12:16:39 Integrated Voltage Measurement... 12:16:48 Voltage Offset Correction: v_off_ini = 729.761 uV, v_off_fin = 734.453 uV 12:16:48 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 248.709 uVs 12:16:48 Integrated Voltage Measurement... 12:16:57 Voltage Offset Correction: v_off_ini = 725.064 uV, v_off_fin = 714.000 uV 12:16:57 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -251.817 uVs 12:16:57 Integrated Voltage Measurement... 12:17:06 Voltage Offset Correction: v_off_ini = 727.714 uV, v_off_fin = 732.279 uV 12:17:06 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 252.734 uVs 12:17:10 I1Y Average and RMS of Measurements VT = -257.189 +- 3.693 uVs I1Y = -51.438 +- 0.739 uTm 12:17:15 Wire Position: X0 = 2.0000 mm, Y0 = 0.0000 mm 12:17:18 Integrated Voltage Measurement... 12:17:27 Voltage Offset Correction: v_off_ini = 720.820 uV, v_off_fin = 694.658 uV 12:17:27 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -72.761 uVs 12:17:27 Integrated Voltage Measurement... 12:17:36 Voltage Offset Correction: v_off_ini = 717.415 uV, v_off_fin = 722.357 uV 12:17:36 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 77.255 uVs 12:17:36 Integrated Voltage Measurement... 12:17:45 Voltage Offset Correction: v_off_ini = 732.236 uV, v_off_fin = 717.970 uV 12:17:45 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -88.570 uVs 12:17:45 Integrated Voltage Measurement... 12:17:54 Voltage Offset Correction: v_off_ini = 720.171 uV, v_off_fin = 723.969 uV 12:17:54 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 112.383 uVs 12:17:54 Integrated Voltage Measurement... 12:18:03 Voltage Offset Correction: v_off_ini = 722.923 uV, v_off_fin = 710.319 uV 12:18:03 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -88.444 uVs 12:18:03 Integrated Voltage Measurement... 12:18:12 Voltage Offset Correction: v_off_ini = 733.297 uV, v_off_fin = 731.598 uV 12:18:12 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 83.949 uVs 12:18:15 I2X Average and RMS of Measurements VT = -87.227 +- 10.423 uVs I2X = 8.025 +- 0.959 uTm^2 12:18:20 Wire Position: X0 = 2.0000 mm, Y0 = 0.0000 mm 12:18:22 Integrated Voltage Measurement... 12:18:31 Voltage Offset Correction: v_off_ini = 726.906 uV, v_off_fin = 705.794 uV 12:18:31 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -332.198 uVs 12:18:31 Integrated Voltage Measurement... 12:18:40 Voltage Offset Correction: v_off_ini = 713.682 uV, v_off_fin = 732.480 uV 12:18:40 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 331.976 uVs 12:18:40 Integrated Voltage Measurement... 12:18:49 Voltage Offset Correction: v_off_ini = 731.390 uV, v_off_fin = 724.801 uV 12:18:49 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = -313.004 uVs 12:18:50 Integrated Voltage Measurement... 12:18:58 Voltage Offset Correction: v_off_ini = 724.338 uV, v_off_fin = 735.443 uV 12:18:59 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 339.591 uVs 12:18:59 Integrated Voltage Measurement... 12:19:08 Voltage Offset Correction: v_off_ini = 718.565 uV, v_off_fin = 722.458 uV 12:19:08 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -346.084 uVs 12:19:08 Integrated Voltage Measurement... 12:19:17 Voltage Offset Correction: v_off_ini = 731.072 uV, v_off_fin = 713.925 uV 12:19:17 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 320.545 uVs 12:19:19 I2Y Average and RMS of Measurements VT = -330.566 +- 3.060 uVs I2Y = -30.413 +- 0.281 uTm^2 12:19:27 Wire Position: X0 = 0.0000 mm, Y0 = -1.0000 mm 12:19:31 Integrated Voltage Measurement... 12:19:40 Voltage Offset Correction: v_off_ini = 729.618 uV, v_off_fin = 740.977 uV 12:19:40 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -90.221 uVs 12:19:40 Integrated Voltage Measurement... 12:19:49 Voltage Offset Correction: v_off_ini = 733.814 uV, v_off_fin = 725.332 uV 12:19:49 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 93.069 uVs 12:19:49 Integrated Voltage Measurement... 12:19:58 Voltage Offset Correction: v_off_ini = 729.304 uV, v_off_fin = 726.119 uV 12:19:58 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -65.909 uVs 12:19:58 Integrated Voltage Measurement... 12:20:07 Voltage Offset Correction: v_off_ini = 740.885 uV, v_off_fin = 740.255 uV 12:20:07 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 66.203 uVs 12:20:07 Integrated Voltage Measurement... 12:20:16 Voltage Offset Correction: v_off_ini = 753.531 uV, v_off_fin = 754.559 uV 12:20:16 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -83.744 uVs 12:20:16 Integrated Voltage Measurement... 12:20:25 Voltage Offset Correction: v_off_ini = 740.201 uV, v_off_fin = 747.365 uV 12:20:25 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 97.742 uVs 12:20:29 I1X Average and RMS of Measurements VT = -82.815 +- 11.856 uVs I1X = 16.563 +- 2.371 uTm 12:20:32 Wire Position: X0 = 0.0000 mm, Y0 = -1.0000 mm 12:20:36 Integrated Voltage Measurement... 12:20:45 Voltage Offset Correction: v_off_ini = 754.902 uV, v_off_fin = 735.699 uV 12:20:45 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -268.966 uVs 12:20:45 Integrated Voltage Measurement... 12:20:54 Voltage Offset Correction: v_off_ini = 760.149 uV, v_off_fin = 778.185 uV 12:20:54 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 269.107 uVs 12:20:54 Integrated Voltage Measurement... 12:21:03 Voltage Offset Correction: v_off_ini = 770.994 uV, v_off_fin = 749.930 uV 12:21:03 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -276.001 uVs 12:21:03 Integrated Voltage Measurement... 12:21:12 Voltage Offset Correction: v_off_ini = 758.822 uV, v_off_fin = 775.046 uV 12:21:12 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 227.424 uVs 12:21:12 Integrated Voltage Measurement... 12:21:21 Voltage Offset Correction: v_off_ini = 736.867 uV, v_off_fin = 749.885 uV 12:21:21 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -257.601 uVs 12:21:21 Integrated Voltage Measurement... 12:21:30 Voltage Offset Correction: v_off_ini = 769.455 uV, v_off_fin = 776.643 uV 12:21:30 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 245.437 uVs 12:21:34 I1Y Average and RMS of Measurements VT = -257.423 +- 8.212 uVs I1Y = -51.485 +- 1.642 uTm 12:21:37 Wire Position: X0 = 0.0000 mm, Y0 = -1.0000 mm 12:21:40 Integrated Voltage Measurement... 12:21:49 Voltage Offset Correction: v_off_ini = 762.425 uV, v_off_fin = 750.095 uV 12:21:49 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -116.325 uVs 12:21:49 Integrated Voltage Measurement... 12:21:58 Voltage Offset Correction: v_off_ini = 754.447 uV, v_off_fin = 771.299 uV 12:21:58 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 149.469 uVs 12:21:58 Integrated Voltage Measurement... 12:22:07 Voltage Offset Correction: v_off_ini = 764.503 uV, v_off_fin = 768.091 uV 12:22:07 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -123.513 uVs 12:22:07 Integrated Voltage Measurement... 12:22:16 Voltage Offset Correction: v_off_ini = 760.360 uV, v_off_fin = 773.830 uV 12:22:16 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 108.278 uVs 12:22:16 Integrated Voltage Measurement... 12:22:25 Voltage Offset Correction: v_off_ini = 772.722 uV, v_off_fin = 753.475 uV 12:22:25 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -123.175 uVs 12:22:25 Integrated Voltage Measurement... 12:22:34 Voltage Offset Correction: v_off_ini = 764.561 uV, v_off_fin = 777.204 uV 12:22:34 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 122.961 uVs 12:22:36 I2X Average and RMS of Measurements VT = -123.953 +- 6.969 uVs I2X = 11.404 +- 0.641 uTm^2 12:22:39 Wire Position: X0 = 0.0000 mm, Y0 = -1.0000 mm 12:22:41 Integrated Voltage Measurement... 12:22:50 Voltage Offset Correction: v_off_ini = 776.136 uV, v_off_fin = 776.235 uV 12:22:50 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -346.577 uVs 12:22:50 Integrated Voltage Measurement... 12:22:59 Voltage Offset Correction: v_off_ini = 782.603 uV, v_off_fin = 777.667 uV 12:22:59 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 341.095 uVs 12:22:59 Integrated Voltage Measurement... 12:23:08 Voltage Offset Correction: v_off_ini = 791.699 uV, v_off_fin = 790.453 uV 12:23:08 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -330.947 uVs 12:23:08 Integrated Voltage Measurement... 12:23:17 Voltage Offset Correction: v_off_ini = 783.628 uV, v_off_fin = 785.841 uV 12:23:17 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 335.141 uVs 12:23:17 Integrated Voltage Measurement... 12:23:26 Voltage Offset Correction: v_off_ini = 786.645 uV, v_off_fin = 794.697 uV 12:23:26 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -341.337 uVs 12:23:26 Integrated Voltage Measurement... 12:23:35 Voltage Offset Correction: v_off_ini = 791.200 uV, v_off_fin = 793.280 uV 12:23:35 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 336.686 uVs 12:23:38 I2Y Average and RMS of Measurements VT = -338.630 +- 4.414 uVs I2Y = -31.154 +- 0.406 uTm^2 12:23:42 Wire Position: X0 = 0.0000 mm, Y0 = -0.5000 mm 12:23:46 Integrated Voltage Measurement... 12:23:55 Voltage Offset Correction: v_off_ini = 804.866 uV, v_off_fin = 793.191 uV 12:23:55 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -79.142 uVs 12:23:55 Integrated Voltage Measurement... 12:24:04 Voltage Offset Correction: v_off_ini = 798.624 uV, v_off_fin = 794.586 uV 12:24:04 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 85.973 uVs 12:24:04 Integrated Voltage Measurement... 12:24:13 Voltage Offset Correction: v_off_ini = 785.835 uV, v_off_fin = 787.339 uV 12:24:13 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -45.331 uVs 12:24:13 Integrated Voltage Measurement... 12:24:22 Voltage Offset Correction: v_off_ini = 789.130 uV, v_off_fin = 807.572 uV 12:24:22 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 99.112 uVs 12:24:22 Integrated Voltage Measurement... 12:24:31 Voltage Offset Correction: v_off_ini = 808.435 uV, v_off_fin = 788.209 uV 12:24:31 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -60.091 uVs 12:24:31 Integrated Voltage Measurement... 12:24:40 Voltage Offset Correction: v_off_ini = 803.613 uV, v_off_fin = 791.373 uV 12:24:40 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 78.357 uVs 12:24:45 I1X Average and RMS of Measurements VT = -74.668 +- 5.712 uVs I1X = 14.934 +- 1.142 uTm 12:24:47 Wire Position: X0 = 0.0000 mm, Y0 = -0.5000 mm 12:24:51 Integrated Voltage Measurement... 12:25:00 Voltage Offset Correction: v_off_ini = 810.475 uV, v_off_fin = 809.349 uV 12:25:00 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -241.780 uVs 12:25:00 Integrated Voltage Measurement... 12:25:09 Voltage Offset Correction: v_off_ini = 804.603 uV, v_off_fin = 787.628 uV 12:25:09 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 258.146 uVs 12:25:09 Integrated Voltage Measurement... 12:25:18 Voltage Offset Correction: v_off_ini = 781.315 uV, v_off_fin = 793.943 uV 12:25:18 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -241.242 uVs 12:25:18 Integrated Voltage Measurement... 12:25:27 Voltage Offset Correction: v_off_ini = 804.709 uV, v_off_fin = 791.964 uV 12:25:27 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 236.277 uVs 12:25:27 Integrated Voltage Measurement... 12:25:36 Voltage Offset Correction: v_off_ini = 793.610 uV, v_off_fin = 780.572 uV 12:25:36 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -258.650 uVs 12:25:36 Integrated Voltage Measurement... 12:25:45 Voltage Offset Correction: v_off_ini = 797.917 uV, v_off_fin = 791.689 uV 12:25:45 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 248.049 uVs 12:25:50 I1Y Average and RMS of Measurements VT = -247.357 +- 6.235 uVs I1Y = -49.471 +- 1.247 uTm 12:25:52 Wire Position: X0 = 0.0000 mm, Y0 = -0.5000 mm 12:25:55 Integrated Voltage Measurement... 12:26:04 Voltage Offset Correction: v_off_ini = 762.259 uV, v_off_fin = 764.850 uV 12:26:04 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -130.474 uVs 12:26:04 Integrated Voltage Measurement... 12:26:13 Voltage Offset Correction: v_off_ini = 773.393 uV, v_off_fin = 791.373 uV 12:26:13 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 79.122 uVs 12:26:13 Integrated Voltage Measurement... 12:26:22 Voltage Offset Correction: v_off_ini = 767.363 uV, v_off_fin = 746.603 uV 12:26:22 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -119.729 uVs 12:26:22 Integrated Voltage Measurement... 12:26:31 Voltage Offset Correction: v_off_ini = 762.720 uV, v_off_fin = 788.375 uV 12:26:31 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 126.446 uVs 12:26:31 Integrated Voltage Measurement... 12:26:40 Voltage Offset Correction: v_off_ini = 794.669 uV, v_off_fin = 773.870 uV 12:26:40 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -112.759 uVs 12:26:40 Integrated Voltage Measurement... 12:26:49 Voltage Offset Correction: v_off_ini = 782.687 uV, v_off_fin = 797.827 uV 12:26:49 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 92.976 uVs 12:26:51 I2X Average and RMS of Measurements VT = -110.251 +- 9.111 uVs I2X = 10.143 +- 0.838 uTm^2 12:26:54 Wire Position: X0 = 0.0000 mm, Y0 = -0.5000 mm 12:26:56 Integrated Voltage Measurement... 12:27:05 Voltage Offset Correction: v_off_ini = 789.739 uV, v_off_fin = 791.752 uV 12:27:05 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -328.881 uVs 12:27:05 Integrated Voltage Measurement... 12:27:14 Voltage Offset Correction: v_off_ini = 788.564 uV, v_off_fin = 793.840 uV 12:27:14 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 299.294 uVs 12:27:15 Integrated Voltage Measurement... 12:27:24 Voltage Offset Correction: v_off_ini = 802.851 uV, v_off_fin = 803.362 uV 12:27:24 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -337.710 uVs 12:27:24 Integrated Voltage Measurement... 12:27:33 Voltage Offset Correction: v_off_ini = 824.026 uV, v_off_fin = 816.411 uV 12:27:33 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 309.382 uVs 12:27:33 Integrated Voltage Measurement... 12:27:42 Voltage Offset Correction: v_off_ini = 829.700 uV, v_off_fin = 829.290 uV 12:27:42 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -315.847 uVs 12:27:42 Integrated Voltage Measurement... 12:27:51 Voltage Offset Correction: v_off_ini = 835.873 uV, v_off_fin = 825.727 uV 12:27:51 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 350.326 uVs 12:27:53 I2Y Average and RMS of Measurements VT = -323.573 +- 7.756 uVs I2Y = -29.769 +- 0.714 uTm^2 12:27:57 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 12:28:01 Integrated Voltage Measurement... 12:28:10 Voltage Offset Correction: v_off_ini = 814.668 uV, v_off_fin = 803.998 uV 12:28:10 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -56.851 uVs 12:28:10 Integrated Voltage Measurement... 12:28:19 Voltage Offset Correction: v_off_ini = 812.152 uV, v_off_fin = 806.978 uV 12:28:19 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 72.360 uVs 12:28:19 Integrated Voltage Measurement... 12:28:28 Voltage Offset Correction: v_off_ini = 824.991 uV, v_off_fin = 803.926 uV 12:28:28 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -66.640 uVs 12:28:28 Integrated Voltage Measurement... 12:28:37 Voltage Offset Correction: v_off_ini = 813.250 uV, v_off_fin = 825.355 uV 12:28:37 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 60.209 uVs 12:28:37 Integrated Voltage Measurement... 12:28:46 Voltage Offset Correction: v_off_ini = 832.801 uV, v_off_fin = 830.765 uV 12:28:46 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -75.433 uVs 12:28:46 Integrated Voltage Measurement... 12:28:55 Voltage Offset Correction: v_off_ini = 823.183 uV, v_off_fin = 847.451 uV 12:28:55 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 77.249 uVs 12:29:00 I1X Average and RMS of Measurements VT = -68.124 +- 5.831 uVs I1X = 13.625 +- 1.166 uTm 12:29:02 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 12:29:06 Integrated Voltage Measurement... 12:29:15 Voltage Offset Correction: v_off_ini = 855.565 uV, v_off_fin = 812.576 uV 12:29:15 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -266.865 uVs 12:29:15 Integrated Voltage Measurement... 12:29:24 Voltage Offset Correction: v_off_ini = 802.988 uV, v_off_fin = 832.959 uV 12:29:24 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 243.093 uVs 12:29:24 Integrated Voltage Measurement... 12:29:33 Voltage Offset Correction: v_off_ini = 839.208 uV, v_off_fin = 815.530 uV 12:29:33 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -254.409 uVs 12:29:33 Integrated Voltage Measurement... 12:29:42 Voltage Offset Correction: v_off_ini = 840.605 uV, v_off_fin = 865.237 uV 12:29:42 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 251.328 uVs 12:29:42 Integrated Voltage Measurement... 12:29:51 Voltage Offset Correction: v_off_ini = 850.395 uV, v_off_fin = 835.107 uV 12:29:51 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -245.872 uVs 12:29:51 Integrated Voltage Measurement... 12:30:00 Voltage Offset Correction: v_off_ini = 846.784 uV, v_off_fin = 866.405 uV 12:30:00 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 239.563 uVs 12:30:05 I1Y Average and RMS of Measurements VT = -250.189 +- 5.352 uVs I1Y = -50.036 +- 1.073 uTm 12:30:07 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 12:30:10 Integrated Voltage Measurement... 12:30:19 Voltage Offset Correction: v_off_ini = 855.776 uV, v_off_fin = 858.981 uV 12:30:19 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -74.671 uVs 12:30:19 Integrated Voltage Measurement... 12:30:28 Voltage Offset Correction: v_off_ini = 856.239 uV, v_off_fin = 869.825 uV 12:30:28 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 106.423 uVs 12:30:28 Integrated Voltage Measurement... 12:30:37 Voltage Offset Correction: v_off_ini = 855.002 uV, v_off_fin = 837.965 uV 12:30:37 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -98.496 uVs 12:30:37 Integrated Voltage Measurement... 12:30:46 Voltage Offset Correction: v_off_ini = 846.519 uV, v_off_fin = 859.383 uV 12:30:46 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 96.364 uVs 12:30:46 Integrated Voltage Measurement... 12:30:55 Voltage Offset Correction: v_off_ini = 876.511 uV, v_off_fin = 863.967 uV 12:30:55 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -119.443 uVs 12:30:55 Integrated Voltage Measurement... 12:31:04 Voltage Offset Correction: v_off_ini = 872.379 uV, v_off_fin = 878.427 uV 12:31:04 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 87.349 uVs 12:31:06 I2X Average and RMS of Measurements VT = -97.124 +- 5.250 uVs I2X = 8.935 +- 0.483 uTm^2 12:31:09 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 12:31:12 Integrated Voltage Measurement... 12:31:21 Voltage Offset Correction: v_off_ini = 857.845 uV, v_off_fin = 854.763 uV 12:31:21 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -321.162 uVs 12:31:21 Integrated Voltage Measurement... 12:31:30 Voltage Offset Correction: v_off_ini = 854.215 uV, v_off_fin = 866.643 uV 12:31:30 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 327.937 uVs 12:31:30 Integrated Voltage Measurement... 12:31:39 Voltage Offset Correction: v_off_ini = 870.464 uV, v_off_fin = 862.707 uV 12:31:39 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -304.502 uVs 12:31:39 Integrated Voltage Measurement... 12:31:48 Voltage Offset Correction: v_off_ini = 868.303 uV, v_off_fin = 856.652 uV 12:31:48 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 342.426 uVs 12:31:48 Integrated Voltage Measurement... 12:31:57 Voltage Offset Correction: v_off_ini = 876.683 uV, v_off_fin = 854.006 uV 12:31:57 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -328.736 uVs 12:31:57 Integrated Voltage Measurement... 12:32:06 Voltage Offset Correction: v_off_ini = 874.504 uV, v_off_fin = 856.603 uV 12:32:06 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 323.843 uVs 12:32:08 I2Y Average and RMS of Measurements VT = -324.768 +- 1.164 uVs I2Y = -29.879 +- 0.107 uTm^2 12:32:12 Wire Position: X0 = 0.0000 mm, Y0 = 0.5000 mm 12:32:16 Integrated Voltage Measurement... 12:32:25 Voltage Offset Correction: v_off_ini = 859.736 uV, v_off_fin = 860.374 uV 12:32:25 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.4998 mm dXY = -0.4999 mm VT = -77.260 uVs 12:32:26 Integrated Voltage Measurement... 12:32:34 Voltage Offset Correction: v_off_ini = 870.118 uV, v_off_fin = 866.353 uV 12:32:35 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 65.266 uVs 12:32:35 Integrated Voltage Measurement... 12:32:43 Voltage Offset Correction: v_off_ini = 861.373 uV, v_off_fin = 848.878 uV 12:32:44 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -50.363 uVs 12:32:44 Integrated Voltage Measurement... 12:32:53 Voltage Offset Correction: v_off_ini = 843.347 uV, v_off_fin = 846.245 uV 12:32:53 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 60.437 uVs 12:32:53 Integrated Voltage Measurement... 12:33:02 Voltage Offset Correction: v_off_ini = 860.939 uV, v_off_fin = 825.410 uV 12:33:02 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -64.789 uVs 12:33:02 Integrated Voltage Measurement... 12:33:11 Voltage Offset Correction: v_off_ini = 810.900 uV, v_off_fin = 824.301 uV 12:33:11 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 64.007 uVs 12:33:15 I1X Average and RMS of Measurements VT = -63.687 +- 6.496 uVs I1X = 12.738 +- 1.300 uTm 12:33:17 Wire Position: X0 = 0.0000 mm, Y0 = 0.5000 mm 12:33:22 Integrated Voltage Measurement... 12:33:31 Voltage Offset Correction: v_off_ini = 826.019 uV, v_off_fin = 823.979 uV 12:33:31 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -230.901 uVs 12:33:31 Integrated Voltage Measurement... 12:33:40 Voltage Offset Correction: v_off_ini = 839.574 uV, v_off_fin = 855.857 uV 12:33:40 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 240.661 uVs 12:33:40 Integrated Voltage Measurement... 12:33:49 Voltage Offset Correction: v_off_ini = 850.007 uV, v_off_fin = 834.803 uV 12:33:49 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -249.012 uVs 12:33:49 Integrated Voltage Measurement... 12:33:58 Voltage Offset Correction: v_off_ini = 834.873 uV, v_off_fin = 842.418 uV 12:33:58 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 248.392 uVs 12:33:58 Integrated Voltage Measurement... 12:34:07 Voltage Offset Correction: v_off_ini = 818.540 uV, v_off_fin = 811.951 uV 12:34:07 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -232.041 uVs 12:34:07 Integrated Voltage Measurement... 12:34:16 Voltage Offset Correction: v_off_ini = 844.275 uV, v_off_fin = 865.283 uV 12:34:16 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 243.816 uVs 12:34:20 I1Y Average and RMS of Measurements VT = -240.804 +- 5.653 uVs I1Y = -48.158 +- 1.131 uTm 12:34:22 Wire Position: X0 = 0.0000 mm, Y0 = 0.5000 mm 12:34:25 Integrated Voltage Measurement... 12:34:34 Voltage Offset Correction: v_off_ini = 846.183 uV, v_off_fin = 833.834 uV 12:34:34 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -72.501 uVs 12:34:34 Integrated Voltage Measurement... 12:34:43 Voltage Offset Correction: v_off_ini = 852.980 uV, v_off_fin = 856.589 uV 12:34:43 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 80.796 uVs 12:34:43 Integrated Voltage Measurement... 12:34:52 Voltage Offset Correction: v_off_ini = 847.512 uV, v_off_fin = 842.484 uV 12:34:52 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -83.104 uVs 12:34:52 Integrated Voltage Measurement... 12:35:01 Voltage Offset Correction: v_off_ini = 821.464 uV, v_off_fin = 838.470 uV 12:35:01 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 93.264 uVs 12:35:01 Integrated Voltage Measurement... 12:35:10 Voltage Offset Correction: v_off_ini = 839.894 uV, v_off_fin = 823.594 uV 12:35:10 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -71.449 uVs 12:35:10 Integrated Voltage Measurement... 12:35:19 Voltage Offset Correction: v_off_ini = 840.783 uV, v_off_fin = 848.794 uV 12:35:19 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 101.618 uVs 12:35:21 I2X Average and RMS of Measurements VT = -83.789 +- 5.094 uVs I2X = 7.709 +- 0.469 uTm^2 12:35:24 Wire Position: X0 = 0.0000 mm, Y0 = 0.5000 mm 12:35:27 Integrated Voltage Measurement... 12:35:36 Voltage Offset Correction: v_off_ini = 833.733 uV, v_off_fin = 849.158 uV 12:35:36 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -317.257 uVs 12:35:36 Integrated Voltage Measurement... 12:35:45 Voltage Offset Correction: v_off_ini = 854.361 uV, v_off_fin = 857.339 uV 12:35:45 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 316.991 uVs 12:35:45 Integrated Voltage Measurement... 12:35:54 Voltage Offset Correction: v_off_ini = 849.205 uV, v_off_fin = 844.354 uV 12:35:54 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -343.362 uVs 12:35:54 Integrated Voltage Measurement... 12:36:03 Voltage Offset Correction: v_off_ini = 845.701 uV, v_off_fin = 836.222 uV 12:36:03 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = 324.547 uVs 12:36:03 Integrated Voltage Measurement... 12:36:12 Voltage Offset Correction: v_off_ini = 833.270 uV, v_off_fin = 838.311 uV 12:36:12 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -354.613 uVs 12:36:12 Integrated Voltage Measurement... 12:36:21 Voltage Offset Correction: v_off_ini = 850.997 uV, v_off_fin = 839.849 uV 12:36:21 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 345.571 uVs 12:36:23 I2Y Average and RMS of Measurements VT = -333.724 +- 13.460 uVs I2Y = -30.703 +- 1.238 uTm^2 12:36:27 Wire Position: X0 = 0.0000 mm, Y0 = 1.0000 mm 12:36:32 Integrated Voltage Measurement... 12:36:41 Voltage Offset Correction: v_off_ini = 844.909 uV, v_off_fin = 864.175 uV 12:36:41 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -50.547 uVs 12:36:41 Integrated Voltage Measurement... 12:36:50 Voltage Offset Correction: v_off_ini = 849.695 uV, v_off_fin = 867.466 uV 12:36:50 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 49.312 uVs 12:36:50 Integrated Voltage Measurement... 12:36:59 Voltage Offset Correction: v_off_ini = 863.404 uV, v_off_fin = 861.016 uV 12:36:59 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -49.628 uVs 12:36:59 Integrated Voltage Measurement... 12:37:08 Voltage Offset Correction: v_off_ini = 864.866 uV, v_off_fin = 862.510 uV 12:37:08 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 52.806 uVs 12:37:08 Integrated Voltage Measurement... 12:37:17 Voltage Offset Correction: v_off_ini = 873.364 uV, v_off_fin = 864.565 uV 12:37:17 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -63.774 uVs 12:37:17 Integrated Voltage Measurement... 12:37:26 Voltage Offset Correction: v_off_ini = 845.757 uV, v_off_fin = 849.367 uV 12:37:26 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 51.754 uVs 12:37:30 I1X Average and RMS of Measurements VT = -52.970 +- 3.430 uVs I1X = 10.594 +- 0.686 uTm 12:37:32 Wire Position: X0 = 0.0000 mm, Y0 = 1.0000 mm 12:37:37 Integrated Voltage Measurement... 12:37:45 Voltage Offset Correction: v_off_ini = 826.200 uV, v_off_fin = 829.203 uV 12:37:45 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -252.511 uVs 12:37:45 Integrated Voltage Measurement... 12:37:54 Voltage Offset Correction: v_off_ini = 841.064 uV, v_off_fin = 861.203 uV 12:37:54 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 261.575 uVs 12:37:54 Integrated Voltage Measurement... 12:38:03 Voltage Offset Correction: v_off_ini = 841.411 uV, v_off_fin = 836.276 uV 12:38:03 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5002 mm dXY = 0.5001 mm VT = -246.285 uVs 12:38:03 Integrated Voltage Measurement... 12:38:12 Voltage Offset Correction: v_off_ini = 856.006 uV, v_off_fin = 859.569 uV 12:38:12 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5002 mm dXY = -0.5001 mm VT = 234.405 uVs 12:38:12 Integrated Voltage Measurement... 12:38:21 Voltage Offset Correction: v_off_ini = 841.755 uV, v_off_fin = 839.893 uV 12:38:21 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5002 mm dXY = 0.5001 mm VT = -238.634 uVs 12:38:21 Integrated Voltage Measurement... 12:38:30 Voltage Offset Correction: v_off_ini = 857.770 uV, v_off_fin = 857.020 uV 12:38:30 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5002 mm dXY = -0.5001 mm VT = 225.900 uVs 12:38:35 I1Y Average and RMS of Measurements VT = -243.218 +- 10.317 uVs I1Y = -48.636 +- 2.066 uTm 12:38:37 Wire Position: X0 = 0.0000 mm, Y0 = 1.0000 mm 12:38:40 Integrated Voltage Measurement... 12:38:49 Voltage Offset Correction: v_off_ini = 848.274 uV, v_off_fin = 824.451 uV 12:38:49 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -85.047 uVs 12:38:49 Integrated Voltage Measurement... 12:38:58 Voltage Offset Correction: v_off_ini = 827.873 uV, v_off_fin = 842.004 uV 12:38:58 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 99.122 uVs 12:38:58 Integrated Voltage Measurement... 12:39:07 Voltage Offset Correction: v_off_ini = 854.434 uV, v_off_fin = 839.551 uV 12:39:07 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -78.669 uVs 12:39:07 Integrated Voltage Measurement... 12:39:16 Voltage Offset Correction: v_off_ini = 848.933 uV, v_off_fin = 839.507 uV 12:39:16 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 69.307 uVs 12:39:16 Integrated Voltage Measurement... 12:39:25 Voltage Offset Correction: v_off_ini = 848.097 uV, v_off_fin = 853.936 uV 12:39:25 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -74.804 uVs 12:39:25 Integrated Voltage Measurement... 12:39:34 Voltage Offset Correction: v_off_ini = 836.807 uV, v_off_fin = 851.199 uV 12:39:34 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 84.302 uVs 12:39:36 I2X Average and RMS of Measurements VT = -81.875 +- 7.568 uVs I2X = 7.533 +- 0.696 uTm^2 12:39:39 Wire Position: X0 = 0.0000 mm, Y0 = 1.0000 mm 12:39:41 Integrated Voltage Measurement... 12:39:50 Voltage Offset Correction: v_off_ini = 862.547 uV, v_off_fin = 838.311 uV 12:39:50 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -331.371 uVs 12:39:50 Integrated Voltage Measurement... 12:39:59 Voltage Offset Correction: v_off_ini = 851.984 uV, v_off_fin = 875.765 uV 12:39:59 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 317.747 uVs 12:39:59 Integrated Voltage Measurement... 12:40:08 Voltage Offset Correction: v_off_ini = 867.994 uV, v_off_fin = 864.960 uV 12:40:08 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -330.987 uVs 12:40:08 Integrated Voltage Measurement... 12:40:17 Voltage Offset Correction: v_off_ini = 853.332 uV, v_off_fin = 863.971 uV 12:40:17 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 323.837 uVs 12:40:17 Integrated Voltage Measurement... 12:40:26 Voltage Offset Correction: v_off_ini = 840.613 uV, v_off_fin = 870.425 uV 12:40:26 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -338.735 uVs 12:40:27 Integrated Voltage Measurement... 12:40:35 Voltage Offset Correction: v_off_ini = 878.349 uV, v_off_fin = 904.697 uV 12:40:36 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 341.763 uVs 12:40:38 I2Y Average and RMS of Measurements VT = -330.740 +- 6.824 uVs I2Y = -30.428 +- 0.628 uTm^2 12:40:54 Wire Position: X0 = -2.0000 mm, Y0 = 0.0000 mm 12:40:58 Integrated Voltage Measurement... 12:41:07 Voltage Offset Correction: v_off_ini = 883.249 uV, v_off_fin = 870.075 uV 12:41:07 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -64.783 uVs 12:41:07 Integrated Voltage Measurement... 12:41:16 Voltage Offset Correction: v_off_ini = 867.886 uV, v_off_fin = 859.387 uV 12:41:16 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 64.241 uVs 12:41:16 Integrated Voltage Measurement... 12:41:25 Voltage Offset Correction: v_off_ini = 882.726 uV, v_off_fin = 887.753 uV 12:41:25 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -60.748 uVs 12:41:25 Integrated Voltage Measurement... 12:41:34 Voltage Offset Correction: v_off_ini = 882.905 uV, v_off_fin = 859.952 uV 12:41:34 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 58.182 uVs 12:41:34 Integrated Voltage Measurement... 12:41:43 Voltage Offset Correction: v_off_ini = 873.691 uV, v_off_fin = 902.454 uV 12:41:43 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -66.345 uVs 12:41:43 Integrated Voltage Measurement... 12:41:52 Voltage Offset Correction: v_off_ini = 882.811 uV, v_off_fin = 876.467 uV 12:41:52 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 65.676 uVs 12:41:56 I1X Average and RMS of Measurements VT = -63.329 +- 2.800 uVs I1X = 12.666 +- 0.560 uTm 12:41:59 Wire Position: X0 = -2.0000 mm, Y0 = 0.0000 mm 12:42:03 Integrated Voltage Measurement... 12:42:12 Voltage Offset Correction: v_off_ini = 858.576 uV, v_off_fin = 864.744 uV 12:42:12 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.4999 mm dXY = 0.4999 mm VT = -253.100 uVs 12:42:12 Integrated Voltage Measurement... 12:42:21 Voltage Offset Correction: v_off_ini = 881.207 uV, v_off_fin = 887.715 uV 12:42:21 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 237.924 uVs 12:42:21 Integrated Voltage Measurement... 12:42:30 Voltage Offset Correction: v_off_ini = 879.364 uV, v_off_fin = 870.128 uV 12:42:30 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -245.196 uVs 12:42:30 Integrated Voltage Measurement... 12:42:39 Voltage Offset Correction: v_off_ini = 873.515 uV, v_off_fin = 895.791 uV 12:42:39 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 252.852 uVs 12:42:39 Integrated Voltage Measurement... 12:42:48 Voltage Offset Correction: v_off_ini = 904.808 uV, v_off_fin = 889.558 uV 12:42:48 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -264.573 uVs 12:42:48 Integrated Voltage Measurement... 12:42:57 Voltage Offset Correction: v_off_ini = 874.521 uV, v_off_fin = 893.676 uV 12:42:57 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 256.459 uVs 12:43:01 I1Y Average and RMS of Measurements VT = -251.684 +- 6.408 uVs I1Y = -50.338 +- 1.280 uTm 12:43:06 Wire Position: X0 = -2.0000 mm, Y0 = 0.0000 mm 12:43:09 Integrated Voltage Measurement... 12:43:18 Voltage Offset Correction: v_off_ini = 908.088 uV, v_off_fin = 896.339 uV 12:43:18 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -78.693 uVs 12:43:18 Integrated Voltage Measurement... 12:43:27 Voltage Offset Correction: v_off_ini = 914.973 uV, v_off_fin = 903.465 uV 12:43:27 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 95.440 uVs 12:43:27 Integrated Voltage Measurement... 12:43:36 Voltage Offset Correction: v_off_ini = 899.765 uV, v_off_fin = 887.158 uV 12:43:36 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -103.008 uVs 12:43:36 Integrated Voltage Measurement... 12:43:45 Voltage Offset Correction: v_off_ini = 889.192 uV, v_off_fin = 907.729 uV 12:43:45 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 97.051 uVs 12:43:45 Integrated Voltage Measurement... 12:43:54 Voltage Offset Correction: v_off_ini = 914.145 uV, v_off_fin = 898.461 uV 12:43:54 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -85.105 uVs 12:43:54 Integrated Voltage Measurement... 12:44:03 Voltage Offset Correction: v_off_ini = 891.904 uV, v_off_fin = 893.261 uV 12:44:03 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 103.174 uVs 12:44:05 I2X Average and RMS of Measurements VT = -93.745 +- 5.300 uVs I2X = 8.625 +- 0.488 uTm^2 12:44:11 Wire Position: X0 = -2.0000 mm, Y0 = 0.0000 mm 12:44:13 Integrated Voltage Measurement... 12:44:22 Voltage Offset Correction: v_off_ini = 930.499 uV, v_off_fin = 915.273 uV 12:44:22 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -347.916 uVs 12:44:22 Integrated Voltage Measurement... 12:44:31 Voltage Offset Correction: v_off_ini = 919.652 uV, v_off_fin = 920.472 uV 12:44:31 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 314.949 uVs 12:44:31 Integrated Voltage Measurement... 12:44:40 Voltage Offset Correction: v_off_ini = 914.302 uV, v_off_fin = 905.765 uV 12:44:40 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -363.666 uVs 12:44:40 Integrated Voltage Measurement... 12:44:49 Voltage Offset Correction: v_off_ini = 897.324 uV, v_off_fin = 882.409 uV 12:44:49 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 318.181 uVs 12:44:49 Integrated Voltage Measurement... 12:44:58 Voltage Offset Correction: v_off_ini = 897.652 uV, v_off_fin = 893.771 uV 12:44:58 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -332.901 uVs 12:44:58 Integrated Voltage Measurement... 12:45:07 Voltage Offset Correction: v_off_ini = 917.608 uV, v_off_fin = 891.528 uV 12:45:07 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 358.914 uVs 12:45:10 I2Y Average and RMS of Measurements VT = -339.421 +- 6.004 uVs I2Y = -31.227 +- 0.552 uTm^2 12:45:14 Wire Position: X0 = -1.0000 mm, Y0 = 0.0000 mm 12:45:19 Integrated Voltage Measurement... 12:45:28 Voltage Offset Correction: v_off_ini = 930.658 uV, v_off_fin = 908.125 uV 12:45:28 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -64.742 uVs 12:45:28 Integrated Voltage Measurement... 12:45:37 Voltage Offset Correction: v_off_ini = 915.209 uV, v_off_fin = 897.414 uV 12:45:37 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 83.768 uVs 12:45:37 Integrated Voltage Measurement... 12:45:46 Voltage Offset Correction: v_off_ini = 910.647 uV, v_off_fin = 923.158 uV 12:45:46 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -74.417 uVs 12:45:46 Integrated Voltage Measurement... 12:45:55 Voltage Offset Correction: v_off_ini = 928.474 uV, v_off_fin = 920.307 uV 12:45:55 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 22.467 uVs 12:45:55 Integrated Voltage Measurement... 12:46:04 Voltage Offset Correction: v_off_ini = 942.069 uV, v_off_fin = 940.003 uV 12:46:04 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -83.711 uVs 12:46:04 Integrated Voltage Measurement... 12:46:13 Voltage Offset Correction: v_off_ini = 917.050 uV, v_off_fin = 923.378 uV 12:46:13 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 67.455 uVs 12:46:17 I1X Average and RMS of Measurements VT = -66.093 +- 12.493 uVs I1X = 13.219 +- 2.499 uTm 12:46:19 Wire Position: X0 = -1.0000 mm, Y0 = 0.0000 mm 12:46:24 Integrated Voltage Measurement... 12:46:33 Voltage Offset Correction: v_off_ini = 916.275 uV, v_off_fin = 908.869 uV 12:46:33 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -253.092 uVs 12:46:33 Integrated Voltage Measurement... 12:46:42 Voltage Offset Correction: v_off_ini = 936.844 uV, v_off_fin = 931.197 uV 12:46:42 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 236.315 uVs 12:46:42 Integrated Voltage Measurement... 12:46:51 Voltage Offset Correction: v_off_ini = 905.431 uV, v_off_fin = 911.467 uV 12:46:51 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -252.375 uVs 12:46:51 Integrated Voltage Measurement... 12:47:00 Voltage Offset Correction: v_off_ini = 907.086 uV, v_off_fin = 913.327 uV 12:47:00 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 258.981 uVs 12:47:00 Integrated Voltage Measurement... 12:47:08 Voltage Offset Correction: v_off_ini = 894.391 uV, v_off_fin = 895.021 uV 12:47:09 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -255.672 uVs 12:47:09 Integrated Voltage Measurement... 12:47:17 Voltage Offset Correction: v_off_ini = 922.886 uV, v_off_fin = 954.034 uV 12:47:18 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 237.346 uVs 12:47:22 I1Y Average and RMS of Measurements VT = -248.963 +- 4.805 uVs I1Y = -49.793 +- 0.961 uTm 12:47:27 Wire Position: X0 = -1.0000 mm, Y0 = 0.0000 mm 12:47:29 Integrated Voltage Measurement... 12:47:38 Voltage Offset Correction: v_off_ini = 923.330 uV, v_off_fin = 917.567 uV 12:47:38 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -87.049 uVs 12:47:38 Integrated Voltage Measurement... 12:47:47 Voltage Offset Correction: v_off_ini = 915.176 uV, v_off_fin = 934.147 uV 12:47:47 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 96.536 uVs 12:47:47 Integrated Voltage Measurement... 12:47:56 Voltage Offset Correction: v_off_ini = 927.577 uV, v_off_fin = 903.647 uV 12:47:56 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -99.186 uVs 12:47:57 Integrated Voltage Measurement... 12:48:05 Voltage Offset Correction: v_off_ini = 929.146 uV, v_off_fin = 953.013 uV 12:48:05 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 79.548 uVs 12:48:06 Integrated Voltage Measurement... 12:48:15 Voltage Offset Correction: v_off_ini = 946.167 uV, v_off_fin = 921.251 uV 12:48:15 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -72.251 uVs 12:48:15 Integrated Voltage Measurement... 12:48:24 Voltage Offset Correction: v_off_ini = 899.675 uV, v_off_fin = 920.120 uV 12:48:24 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 87.020 uVs 12:48:26 I2X Average and RMS of Measurements VT = -86.932 +- 5.253 uVs I2X = 7.998 +- 0.483 uTm^2 12:48:31 Wire Position: X0 = -1.0000 mm, Y0 = 0.0000 mm 12:48:34 Integrated Voltage Measurement... 12:48:43 Voltage Offset Correction: v_off_ini = 925.747 uV, v_off_fin = 933.333 uV 12:48:43 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -336.124 uVs 12:48:43 Integrated Voltage Measurement... 12:48:52 Voltage Offset Correction: v_off_ini = 937.589 uV, v_off_fin = 932.770 uV 12:48:52 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 340.426 uVs 12:48:52 Integrated Voltage Measurement... 12:49:01 Voltage Offset Correction: v_off_ini = 924.359 uV, v_off_fin = 928.853 uV 12:49:01 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -346.887 uVs 12:49:01 Integrated Voltage Measurement... 12:49:10 Voltage Offset Correction: v_off_ini = 935.599 uV, v_off_fin = 947.792 uV 12:49:10 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 329.481 uVs 12:49:10 Integrated Voltage Measurement... 12:49:19 Voltage Offset Correction: v_off_ini = 944.472 uV, v_off_fin = 939.730 uV 12:49:19 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -329.196 uVs 12:49:19 Integrated Voltage Measurement... 12:49:28 Voltage Offset Correction: v_off_ini = 941.368 uV, v_off_fin = 940.529 uV 12:49:28 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 347.964 uVs 12:49:30 I2Y Average and RMS of Measurements VT = -338.346 +- 0.169 uVs I2Y = -31.128 +- 0.016 uTm^2 12:49:35 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 12:49:39 Integrated Voltage Measurement... 12:49:48 Voltage Offset Correction: v_off_ini = 932.768 uV, v_off_fin = 921.199 uV 12:49:48 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -74.347 uVs 12:49:48 Integrated Voltage Measurement... 12:49:57 Voltage Offset Correction: v_off_ini = 934.220 uV, v_off_fin = 937.594 uV 12:49:57 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 61.219 uVs 12:49:57 Integrated Voltage Measurement... 12:50:06 Voltage Offset Correction: v_off_ini = 933.043 uV, v_off_fin = 942.760 uV 12:50:06 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -69.544 uVs 12:50:06 Integrated Voltage Measurement... 12:50:15 Voltage Offset Correction: v_off_ini = 961.146 uV, v_off_fin = 955.295 uV 12:50:15 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 26.305 uVs 12:50:15 Integrated Voltage Measurement... 12:50:24 Voltage Offset Correction: v_off_ini = 952.947 uV, v_off_fin = 954.846 uV 12:50:24 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -55.071 uVs 12:50:24 Integrated Voltage Measurement... 12:50:33 Voltage Offset Correction: v_off_ini = 923.866 uV, v_off_fin = 930.554 uV 12:50:33 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 56.933 uVs 12:50:37 I1X Average and RMS of Measurements VT = -57.236 +- 8.154 uVs I1X = 11.448 +- 1.631 uTm 12:50:40 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 12:50:44 Integrated Voltage Measurement... 12:50:53 Voltage Offset Correction: v_off_ini = 941.442 uV, v_off_fin = 930.085 uV 12:50:53 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = -262.045 uVs 12:50:53 Integrated Voltage Measurement... 12:51:02 Voltage Offset Correction: v_off_ini = 937.212 uV, v_off_fin = 968.740 uV 12:51:02 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 258.000 uVs 12:51:02 Integrated Voltage Measurement... 12:51:11 Voltage Offset Correction: v_off_ini = 932.788 uV, v_off_fin = 922.808 uV 12:51:11 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -247.849 uVs 12:51:11 Integrated Voltage Measurement... 12:51:20 Voltage Offset Correction: v_off_ini = 929.869 uV, v_off_fin = 967.047 uV 12:51:20 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 246.323 uVs 12:51:20 Integrated Voltage Measurement... 12:51:29 Voltage Offset Correction: v_off_ini = 953.623 uV, v_off_fin = 940.165 uV 12:51:29 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -231.422 uVs 12:51:29 Integrated Voltage Measurement... 12:51:38 Voltage Offset Correction: v_off_ini = 938.349 uV, v_off_fin = 957.870 uV 12:51:38 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 263.764 uVs 12:51:42 I1Y Average and RMS of Measurements VT = -251.567 +- 5.982 uVs I1Y = -50.314 +- 1.198 uTm 12:51:48 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 12:51:50 Integrated Voltage Measurement... 12:51:59 Voltage Offset Correction: v_off_ini = 951.425 uV, v_off_fin = 952.010 uV 12:51:59 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -85.457 uVs 12:51:59 Integrated Voltage Measurement... 12:52:08 Voltage Offset Correction: v_off_ini = 950.702 uV, v_off_fin = 974.103 uV 12:52:08 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 72.191 uVs 12:52:08 Integrated Voltage Measurement... 12:52:17 Voltage Offset Correction: v_off_ini = 977.628 uV, v_off_fin = 961.903 uV 12:52:17 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -81.711 uVs 12:52:17 Integrated Voltage Measurement... 12:52:26 Voltage Offset Correction: v_off_ini = 949.813 uV, v_off_fin = 944.496 uV 12:52:26 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 92.545 uVs 12:52:26 Integrated Voltage Measurement... 12:52:35 Voltage Offset Correction: v_off_ini = 963.746 uV, v_off_fin = 962.388 uV 12:52:35 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -86.438 uVs 12:52:35 Integrated Voltage Measurement... 12:52:44 Voltage Offset Correction: v_off_ini = 963.872 uV, v_off_fin = 966.774 uV 12:52:44 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 84.130 uVs 12:52:47 I2X Average and RMS of Measurements VT = -83.745 +- 3.560 uVs I2X = 7.705 +- 0.328 uTm^2 12:52:52 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 12:52:54 Integrated Voltage Measurement... 12:53:03 Voltage Offset Correction: v_off_ini = 975.397 uV, v_off_fin = 964.773 uV 12:53:03 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -326.978 uVs 12:53:03 Integrated Voltage Measurement... 12:53:12 Voltage Offset Correction: v_off_ini = 965.409 uV, v_off_fin = 950.426 uV 12:53:12 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 375.407 uVs 12:53:13 Integrated Voltage Measurement... 12:53:21 Voltage Offset Correction: v_off_ini = 965.784 uV, v_off_fin = 961.065 uV 12:53:22 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -316.858 uVs 12:53:22 Integrated Voltage Measurement... 12:53:31 Voltage Offset Correction: v_off_ini = 965.965 uV, v_off_fin = 980.669 uV 12:53:31 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 334.182 uVs 12:53:31 Integrated Voltage Measurement... 12:53:40 Voltage Offset Correction: v_off_ini = 974.613 uV, v_off_fin = 949.545 uV 12:53:40 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -319.094 uVs 12:53:40 Integrated Voltage Measurement... 12:53:49 Voltage Offset Correction: v_off_ini = 959.817 uV, v_off_fin = 980.454 uV 12:53:49 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 334.674 uVs 12:53:51 I2Y Average and RMS of Measurements VT = -334.532 +- 11.794 uVs I2Y = -30.777 +- 1.085 uTm^2 12:53:56 Wire Position: X0 = 1.0000 mm, Y0 = 0.0000 mm 12:54:00 Integrated Voltage Measurement... 12:54:09 Voltage Offset Correction: v_off_ini = 970.989 uV, v_off_fin = 953.308 uV 12:54:09 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -71.779 uVs 12:54:09 Integrated Voltage Measurement... 12:54:18 Voltage Offset Correction: v_off_ini = 952.882 uV, v_off_fin = 957.430 uV 12:54:18 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 101.124 uVs 12:54:18 Integrated Voltage Measurement... 12:54:27 Voltage Offset Correction: v_off_ini = 951.855 uV, v_off_fin = 957.706 uV 12:54:27 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -83.493 uVs 12:54:27 Integrated Voltage Measurement... 12:54:36 Voltage Offset Correction: v_off_ini = 957.555 uV, v_off_fin = 973.048 uV 12:54:36 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 72.871 uVs 12:54:36 Integrated Voltage Measurement... 12:54:45 Voltage Offset Correction: v_off_ini = 956.158 uV, v_off_fin = 980.748 uV 12:54:45 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -66.501 uVs 12:54:45 Integrated Voltage Measurement... 12:54:54 Voltage Offset Correction: v_off_ini = 964.379 uV, v_off_fin = 976.547 uV 12:54:54 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 81.916 uVs 12:54:58 I1X Average and RMS of Measurements VT = -79.614 +- 5.100 uVs I1X = 15.923 +- 1.020 uTm 12:55:01 Wire Position: X0 = 1.0000 mm, Y0 = 0.0000 mm 12:55:05 Integrated Voltage Measurement... 12:55:14 Voltage Offset Correction: v_off_ini = 980.397 uV, v_off_fin = 962.281 uV 12:55:14 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -251.283 uVs 12:55:14 Integrated Voltage Measurement... 12:55:23 Voltage Offset Correction: v_off_ini = 952.592 uV, v_off_fin = 965.148 uV 12:55:23 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 252.499 uVs 12:55:23 Integrated Voltage Measurement... 12:55:32 Voltage Offset Correction: v_off_ini = 966.574 uV, v_off_fin = 958.498 uV 12:55:32 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -264.277 uVs 12:55:32 Integrated Voltage Measurement... 12:55:41 Voltage Offset Correction: v_off_ini = 972.507 uV, v_off_fin = 990.486 uV 12:55:41 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 257.532 uVs 12:55:41 Integrated Voltage Measurement... 12:55:50 Voltage Offset Correction: v_off_ini = 971.977 uV, v_off_fin = 955.938 uV 12:55:50 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -256.829 uVs 12:55:50 Integrated Voltage Measurement... 12:55:59 Voltage Offset Correction: v_off_ini = 958.328 uV, v_off_fin = 966.841 uV 12:55:59 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 256.442 uVs 12:56:03 I1Y Average and RMS of Measurements VT = -256.477 +- 3.681 uVs I1Y = -51.295 +- 0.736 uTm 12:56:08 Wire Position: X0 = 1.0000 mm, Y0 = 0.0000 mm 12:56:11 Integrated Voltage Measurement... 12:56:20 Voltage Offset Correction: v_off_ini = 957.503 uV, v_off_fin = 970.697 uV 12:56:20 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -115.508 uVs 12:56:20 Integrated Voltage Measurement... 12:56:29 Voltage Offset Correction: v_off_ini = 963.092 uV, v_off_fin = 968.162 uV 12:56:29 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 92.996 uVs 12:56:29 Integrated Voltage Measurement... 12:56:38 Voltage Offset Correction: v_off_ini = 945.002 uV, v_off_fin = 962.931 uV 12:56:38 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -79.520 uVs 12:56:38 Integrated Voltage Measurement... 12:56:47 Voltage Offset Correction: v_off_ini = 958.254 uV, v_off_fin = 978.799 uV 12:56:47 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 86.191 uVs 12:56:47 Integrated Voltage Measurement... 12:56:56 Voltage Offset Correction: v_off_ini = 981.331 uV, v_off_fin = 985.502 uV 12:56:56 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -96.073 uVs 12:56:56 Integrated Voltage Measurement... 12:57:05 Voltage Offset Correction: v_off_ini = 967.783 uV, v_off_fin = 971.115 uV 12:57:05 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 108.534 uVs 12:57:08 I2X Average and RMS of Measurements VT = -96.470 +- 9.660 uVs I2X = 8.875 +- 0.889 uTm^2 12:57:13 Wire Position: X0 = 1.0000 mm, Y0 = 0.0000 mm 12:57:15 Integrated Voltage Measurement... 12:57:24 Voltage Offset Correction: v_off_ini = 972.362 uV, v_off_fin = 982.603 uV 12:57:24 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -322.598 uVs 12:57:24 Integrated Voltage Measurement... 12:57:33 Voltage Offset Correction: v_off_ini = 983.729 uV, v_off_fin = 995.010 uV 12:57:33 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 345.826 uVs 12:57:33 Integrated Voltage Measurement... 12:57:42 Voltage Offset Correction: v_off_ini = 982.046 uV, v_off_fin = 997.027 uV 12:57:42 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -327.123 uVs 12:57:42 Integrated Voltage Measurement... 12:57:51 Voltage Offset Correction: v_off_ini = 987.665 uV, v_off_fin = 992.270 uV 12:57:51 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 331.064 uVs 12:57:51 Integrated Voltage Measurement... 12:58:00 Voltage Offset Correction: v_off_ini = 995.689 uV, v_off_fin = 976.922 uV 12:58:00 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -311.622 uVs 12:58:01 Integrated Voltage Measurement... 12:58:09 Voltage Offset Correction: v_off_ini = 943.148 uV, v_off_fin = 957.677 uV 12:58:10 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 347.476 uVs 12:58:12 I2Y Average and RMS of Measurements VT = -330.951 +- 2.313 uVs I2Y = -30.448 +- 0.213 uTm^2 12:58:17 Wire Position: X0 = 2.0000 mm, Y0 = 0.0000 mm 12:58:21 Integrated Voltage Measurement... 12:58:30 Voltage Offset Correction: v_off_ini = 973.497 uV, v_off_fin = 978.720 uV 12:58:30 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -75.916 uVs 12:58:30 Integrated Voltage Measurement... 12:58:39 Voltage Offset Correction: v_off_ini = 980.162 uV, v_off_fin = 958.248 uV 12:58:39 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 73.943 uVs 12:58:39 Integrated Voltage Measurement... 12:58:48 Voltage Offset Correction: v_off_ini = 952.850 uV, v_off_fin = 963.163 uV 12:58:48 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -93.742 uVs 12:58:48 Integrated Voltage Measurement... 12:58:57 Voltage Offset Correction: v_off_ini = 946.957 uV, v_off_fin = 947.067 uV 12:58:57 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 48.544 uVs 12:58:57 Integrated Voltage Measurement... 12:59:06 Voltage Offset Correction: v_off_ini = 969.197 uV, v_off_fin = 964.086 uV 12:59:06 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -53.184 uVs 12:59:06 Integrated Voltage Measurement... 12:59:15 Voltage Offset Correction: v_off_ini = 956.583 uV, v_off_fin = 953.799 uV 12:59:15 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 85.682 uVs 12:59:19 I1X Average and RMS of Measurements VT = -71.835 +- 2.297 uVs I1X = 14.367 +- 0.459 uTm 12:59:22 Wire Position: X0 = 2.0000 mm, Y0 = 0.0000 mm 12:59:26 Integrated Voltage Measurement... 12:59:35 Voltage Offset Correction: v_off_ini = 969.802 uV, v_off_fin = 935.972 uV 12:59:35 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -251.999 uVs 12:59:35 Integrated Voltage Measurement... 12:59:44 Voltage Offset Correction: v_off_ini = 966.959 uV, v_off_fin = 977.071 uV 12:59:44 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 244.487 uVs 12:59:44 Integrated Voltage Measurement... 12:59:53 Voltage Offset Correction: v_off_ini = 949.572 uV, v_off_fin = 937.745 uV 12:59:53 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -244.465 uVs 12:59:53 Integrated Voltage Measurement... 13:00:02 Voltage Offset Correction: v_off_ini = 937.523 uV, v_off_fin = 959.325 uV 13:00:02 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 250.096 uVs 13:00:02 Integrated Voltage Measurement... 13:00:11 Voltage Offset Correction: v_off_ini = 961.407 uV, v_off_fin = 939.831 uV 13:00:11 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -263.308 uVs 13:00:11 Integrated Voltage Measurement... 13:00:20 Voltage Offset Correction: v_off_ini = 956.844 uV, v_off_fin = 979.698 uV 13:00:20 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 249.900 uVs 13:00:24 I1Y Average and RMS of Measurements VT = -250.709 +- 4.187 uVs I1Y = -50.142 +- 0.837 uTm 13:00:29 Wire Position: X0 = 2.0000 mm, Y0 = 0.0000 mm 13:00:32 Integrated Voltage Measurement... 13:00:41 Voltage Offset Correction: v_off_ini = 960.471 uV, v_off_fin = 961.409 uV 13:00:41 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -90.056 uVs 13:00:41 Integrated Voltage Measurement... 13:00:50 Voltage Offset Correction: v_off_ini = 969.320 uV, v_off_fin = 957.239 uV 13:00:50 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 84.310 uVs 13:00:50 Integrated Voltage Measurement... 13:00:59 Voltage Offset Correction: v_off_ini = 924.682 uV, v_off_fin = 935.170 uV 13:00:59 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -92.268 uVs 13:00:59 Integrated Voltage Measurement... 13:01:08 Voltage Offset Correction: v_off_ini = 942.205 uV, v_off_fin = 951.418 uV 13:01:08 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 86.846 uVs 13:01:08 Integrated Voltage Measurement... 13:01:17 Voltage Offset Correction: v_off_ini = 956.568 uV, v_off_fin = 949.386 uV 13:01:17 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -89.922 uVs 13:01:17 Integrated Voltage Measurement... 13:01:26 Voltage Offset Correction: v_off_ini = 960.522 uV, v_off_fin = 972.096 uV 13:01:26 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 96.110 uVs 13:01:28 I2X Average and RMS of Measurements VT = -89.919 +- 2.395 uVs I2X = 8.273 +- 0.220 uTm^2 13:01:34 Wire Position: X0 = 2.0000 mm, Y0 = 0.0000 mm 13:01:36 Integrated Voltage Measurement... 13:01:45 Voltage Offset Correction: v_off_ini = 918.789 uV, v_off_fin = 934.003 uV 13:01:45 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -310.743 uVs 13:01:45 Integrated Voltage Measurement... 13:01:54 Voltage Offset Correction: v_off_ini = 924.543 uV, v_off_fin = 957.749 uV 13:01:54 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 323.486 uVs 13:01:54 Integrated Voltage Measurement... 13:02:03 Voltage Offset Correction: v_off_ini = 958.486 uV, v_off_fin = 951.686 uV 13:02:03 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -316.462 uVs 13:02:03 Integrated Voltage Measurement... 13:02:12 Voltage Offset Correction: v_off_ini = 977.798 uV, v_off_fin = 965.731 uV 13:02:12 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 322.970 uVs 13:02:12 Integrated Voltage Measurement... 13:02:21 Voltage Offset Correction: v_off_ini = 951.870 uV, v_off_fin = 951.432 uV 13:02:21 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -339.657 uVs 13:02:21 Integrated Voltage Measurement... 13:02:30 Voltage Offset Correction: v_off_ini = 950.642 uV, v_off_fin = 939.120 uV 13:02:30 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 327.910 uVs 13:02:33 I2Y Average and RMS of Measurements VT = -323.538 +- 7.322 uVs I2Y = -29.765 +- 0.674 uTm^2 13:02:41 Wire Position: X0 = 0.0000 mm, Y0 = -1.0000 mm 13:02:45 Integrated Voltage Measurement... 13:02:54 Voltage Offset Correction: v_off_ini = 963.403 uV, v_off_fin = 949.948 uV 13:02:54 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -79.457 uVs 13:02:54 Integrated Voltage Measurement... 13:03:03 Voltage Offset Correction: v_off_ini = 933.907 uV, v_off_fin = 946.434 uV 13:03:03 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 96.947 uVs 13:03:03 Integrated Voltage Measurement... 13:03:12 Voltage Offset Correction: v_off_ini = 964.586 uV, v_off_fin = 954.906 uV 13:03:12 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -91.640 uVs 13:03:12 Integrated Voltage Measurement... 13:03:21 Voltage Offset Correction: v_off_ini = 939.694 uV, v_off_fin = 933.586 uV 13:03:21 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 121.132 uVs 13:03:21 Integrated Voltage Measurement... 13:03:30 Voltage Offset Correction: v_off_ini = 960.490 uV, v_off_fin = 973.848 uV 13:03:30 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -76.723 uVs 13:03:30 Integrated Voltage Measurement... 13:03:39 Voltage Offset Correction: v_off_ini = 949.578 uV, v_off_fin = 954.990 uV 13:03:39 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 92.849 uVs 13:03:43 I1X Average and RMS of Measurements VT = -93.124 +- 9.480 uVs I1X = 18.625 +- 1.896 uTm 13:03:46 Wire Position: X0 = 0.0000 mm, Y0 = -1.0000 mm 13:03:50 Integrated Voltage Measurement... 13:03:59 Voltage Offset Correction: v_off_ini = 961.531 uV, v_off_fin = 938.756 uV 13:03:59 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -277.152 uVs 13:03:59 Integrated Voltage Measurement... 13:04:08 Voltage Offset Correction: v_off_ini = 963.568 uV, v_off_fin = 960.028 uV 13:04:08 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 258.232 uVs 13:04:08 Integrated Voltage Measurement... 13:04:17 Voltage Offset Correction: v_off_ini = 936.034 uV, v_off_fin = 937.335 uV 13:04:17 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -266.303 uVs 13:04:17 Integrated Voltage Measurement... 13:04:26 Voltage Offset Correction: v_off_ini = 933.419 uV, v_off_fin = 936.275 uV 13:04:26 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 245.505 uVs 13:04:26 Integrated Voltage Measurement... 13:04:35 Voltage Offset Correction: v_off_ini = 936.956 uV, v_off_fin = 920.296 uV 13:04:35 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -249.515 uVs 13:04:35 Integrated Voltage Measurement... 13:04:44 Voltage Offset Correction: v_off_ini = 951.356 uV, v_off_fin = 975.053 uV 13:04:44 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 253.928 uVs 13:04:48 I1Y Average and RMS of Measurements VT = -258.439 +- 6.762 uVs I1Y = -51.688 +- 1.352 uTm 13:04:51 Wire Position: X0 = 0.0000 mm, Y0 = -1.0000 mm 13:04:53 Integrated Voltage Measurement... 13:05:02 Voltage Offset Correction: v_off_ini = 954.811 uV, v_off_fin = 941.503 uV 13:05:02 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -108.655 uVs 13:05:02 Integrated Voltage Measurement... 13:05:11 Voltage Offset Correction: v_off_ini = 958.996 uV, v_off_fin = 960.011 uV 13:05:11 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 111.055 uVs 13:05:11 Integrated Voltage Measurement... 13:05:20 Voltage Offset Correction: v_off_ini = 970.417 uV, v_off_fin = 954.520 uV 13:05:20 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -99.496 uVs 13:05:20 Integrated Voltage Measurement... 13:05:29 Voltage Offset Correction: v_off_ini = 970.285 uV, v_off_fin = 964.826 uV 13:05:29 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 101.800 uVs 13:05:29 Integrated Voltage Measurement... 13:05:38 Voltage Offset Correction: v_off_ini = 947.697 uV, v_off_fin = 967.265 uV 13:05:38 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -79.638 uVs 13:05:39 Integrated Voltage Measurement... 13:05:48 Voltage Offset Correction: v_off_ini = 956.189 uV, v_off_fin = 950.893 uV 13:05:48 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 124.203 uVs 13:05:50 I2X Average and RMS of Measurements VT = -104.141 +- 4.074 uVs I2X = 9.581 +- 0.375 uTm^2 13:05:52 Wire Position: X0 = 0.0000 mm, Y0 = -1.0000 mm 13:05:55 Integrated Voltage Measurement... 13:06:04 Voltage Offset Correction: v_off_ini = 962.710 uV, v_off_fin = 968.756 uV 13:06:04 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -339.644 uVs 13:06:04 Integrated Voltage Measurement... 13:06:13 Voltage Offset Correction: v_off_ini = 973.318 uV, v_off_fin = 981.633 uV 13:06:13 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 342.324 uVs 13:06:13 Integrated Voltage Measurement... 13:06:22 Voltage Offset Correction: v_off_ini = 954.638 uV, v_off_fin = 960.533 uV 13:06:22 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -341.577 uVs 13:06:22 Integrated Voltage Measurement... 13:06:31 Voltage Offset Correction: v_off_ini = 945.690 uV, v_off_fin = 955.818 uV 13:06:31 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 360.665 uVs 13:06:31 Integrated Voltage Measurement... 13:06:40 Voltage Offset Correction: v_off_ini = 962.461 uV, v_off_fin = 963.909 uV 13:06:40 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -344.390 uVs 13:06:40 Integrated Voltage Measurement... 13:06:49 Voltage Offset Correction: v_off_ini = 952.774 uV, v_off_fin = 964.505 uV 13:06:49 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 364.086 uVs 13:06:52 I2Y Average and RMS of Measurements VT = -348.781 +- 5.658 uVs I2Y = -32.088 +- 0.521 uTm^2 13:06:56 Wire Position: X0 = 0.0000 mm, Y0 = -0.5000 mm 13:07:00 Integrated Voltage Measurement... 13:07:09 Voltage Offset Correction: v_off_ini = 986.130 uV, v_off_fin = 966.481 uV 13:07:09 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -82.616 uVs 13:07:09 Integrated Voltage Measurement... 13:07:18 Voltage Offset Correction: v_off_ini = 979.074 uV, v_off_fin = 971.737 uV 13:07:18 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 66.443 uVs 13:07:18 Integrated Voltage Measurement... 13:07:27 Voltage Offset Correction: v_off_ini = 974.973 uV, v_off_fin = 968.602 uV 13:07:27 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -67.525 uVs 13:07:27 Integrated Voltage Measurement... 13:07:36 Voltage Offset Correction: v_off_ini = 967.691 uV, v_off_fin = 972.635 uV 13:07:36 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 75.465 uVs 13:07:36 Integrated Voltage Measurement... 13:07:45 Voltage Offset Correction: v_off_ini = 982.215 uV, v_off_fin = 968.559 uV 13:07:45 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -81.985 uVs 13:07:45 Integrated Voltage Measurement... 13:07:54 Voltage Offset Correction: v_off_ini = 962.440 uV, v_off_fin = 953.250 uV 13:07:54 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 84.354 uVs 13:07:58 I1X Average and RMS of Measurements VT = -76.398 +- 4.946 uVs I1X = 15.280 +- 0.989 uTm 13:08:01 Wire Position: X0 = 0.0000 mm, Y0 = -0.5000 mm 13:08:05 Integrated Voltage Measurement... 13:08:14 Voltage Offset Correction: v_off_ini = 972.028 uV, v_off_fin = 978.585 uV 13:08:14 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -260.704 uVs 13:08:14 Integrated Voltage Measurement... 13:08:23 Voltage Offset Correction: v_off_ini = 959.873 uV, v_off_fin = 983.529 uV 13:08:23 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 240.387 uVs 13:08:23 Integrated Voltage Measurement... 13:08:32 Voltage Offset Correction: v_off_ini = 983.491 uV, v_off_fin = 973.398 uV 13:08:32 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -236.641 uVs 13:08:32 Integrated Voltage Measurement... 13:08:41 Voltage Offset Correction: v_off_ini = 984.240 uV, v_off_fin = 991.795 uV 13:08:41 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 249.085 uVs 13:08:41 Integrated Voltage Measurement... 13:08:50 Voltage Offset Correction: v_off_ini = 979.959 uV, v_off_fin = 992.848 uV 13:08:50 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -247.061 uVs 13:08:50 Integrated Voltage Measurement... 13:08:59 Voltage Offset Correction: v_off_ini = 996.913 uV, v_off_fin = 1009.549 uV 13:08:59 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 237.364 uVs 13:09:03 I1Y Average and RMS of Measurements VT = -245.207 +- 3.784 uVs I1Y = -49.041 +- 0.757 uTm 13:09:06 Wire Position: X0 = 0.0000 mm, Y0 = -0.5000 mm 13:09:09 Integrated Voltage Measurement... 13:09:18 Voltage Offset Correction: v_off_ini = 988.142 uV, v_off_fin = 990.113 uV 13:09:18 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -122.380 uVs 13:09:18 Integrated Voltage Measurement... 13:09:27 Voltage Offset Correction: v_off_ini = 990.369 uV, v_off_fin = 1001.110 uV 13:09:27 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 100.736 uVs 13:09:27 Integrated Voltage Measurement... 13:09:36 Voltage Offset Correction: v_off_ini = 1001.508 uV, v_off_fin = 979.123 uV 13:09:36 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -74.582 uVs 13:09:36 Integrated Voltage Measurement... 13:09:45 Voltage Offset Correction: v_off_ini = 989.615 uV, v_off_fin = 1002.134 uV 13:09:45 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 77.986 uVs 13:09:45 Integrated Voltage Measurement... 13:09:54 Voltage Offset Correction: v_off_ini = 990.697 uV, v_off_fin = 999.770 uV 13:09:54 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -98.306 uVs 13:09:54 Integrated Voltage Measurement... 13:10:03 Voltage Offset Correction: v_off_ini = 992.334 uV, v_off_fin = 1012.231 uV 13:10:03 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 129.763 uVs 13:10:05 I2X Average and RMS of Measurements VT = -100.625 +- 17.242 uVs I2X = 9.258 +- 1.586 uTm^2 13:10:08 Wire Position: X0 = 0.0000 mm, Y0 = -0.5000 mm 13:10:10 Integrated Voltage Measurement... 13:10:19 Voltage Offset Correction: v_off_ini = 980.320 uV, v_off_fin = 990.116 uV 13:10:19 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -329.260 uVs 13:10:19 Integrated Voltage Measurement... 13:10:28 Voltage Offset Correction: v_off_ini = 983.566 uV, v_off_fin = 985.630 uV 13:10:28 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 338.040 uVs 13:10:28 Integrated Voltage Measurement... 13:10:37 Voltage Offset Correction: v_off_ini = 972.288 uV, v_off_fin = 983.402 uV 13:10:37 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -340.647 uVs 13:10:37 Integrated Voltage Measurement... 13:10:46 Voltage Offset Correction: v_off_ini = 982.327 uV, v_off_fin = 993.175 uV 13:10:46 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 325.904 uVs 13:10:46 Integrated Voltage Measurement... 13:10:55 Voltage Offset Correction: v_off_ini = 988.595 uV, v_off_fin = 999.539 uV 13:10:55 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -336.801 uVs 13:10:56 Integrated Voltage Measurement... 13:11:04 Voltage Offset Correction: v_off_ini = 965.438 uV, v_off_fin = 981.315 uV 13:11:05 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 340.672 uVs 13:11:07 I2Y Average and RMS of Measurements VT = -335.221 +- 2.491 uVs I2Y = -30.840 +- 0.229 uTm^2 13:11:11 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 13:11:15 Integrated Voltage Measurement... 13:11:24 Voltage Offset Correction: v_off_ini = 970.685 uV, v_off_fin = 978.740 uV 13:11:24 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -73.168 uVs 13:11:24 Integrated Voltage Measurement... 13:11:33 Voltage Offset Correction: v_off_ini = 984.556 uV, v_off_fin = 980.639 uV 13:11:33 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 34.989 uVs 13:11:33 Integrated Voltage Measurement... 13:11:42 Voltage Offset Correction: v_off_ini = 964.094 uV, v_off_fin = 964.098 uV 13:11:42 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -79.599 uVs 13:11:42 Integrated Voltage Measurement... 13:11:51 Voltage Offset Correction: v_off_ini = 974.708 uV, v_off_fin = 976.576 uV 13:11:51 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 81.523 uVs 13:11:51 Integrated Voltage Measurement... 13:12:00 Voltage Offset Correction: v_off_ini = 997.689 uV, v_off_fin = 988.012 uV 13:12:00 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -89.939 uVs 13:12:00 Integrated Voltage Measurement... 13:12:09 Voltage Offset Correction: v_off_ini = 982.750 uV, v_off_fin = 960.336 uV 13:12:09 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 105.224 uVs 13:12:14 I1X Average and RMS of Measurements VT = -77.407 +- 17.899 uVs I1X = 15.481 +- 3.580 uTm 13:12:16 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 13:12:20 Integrated Voltage Measurement... 13:12:29 Voltage Offset Correction: v_off_ini = 978.851 uV, v_off_fin = 983.280 uV 13:12:29 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -250.832 uVs 13:12:29 Integrated Voltage Measurement... 13:12:38 Voltage Offset Correction: v_off_ini = 997.338 uV, v_off_fin = 1010.670 uV 13:12:38 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 231.580 uVs 13:12:38 Integrated Voltage Measurement... 13:12:47 Voltage Offset Correction: v_off_ini = 1013.607 uV, v_off_fin = 976.927 uV 13:12:47 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -279.029 uVs 13:12:47 Integrated Voltage Measurement... 13:12:56 Voltage Offset Correction: v_off_ini = 1000.952 uV, v_off_fin = 996.802 uV 13:12:56 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 228.788 uVs 13:12:56 Integrated Voltage Measurement... 13:13:05 Voltage Offset Correction: v_off_ini = 979.676 uV, v_off_fin = 984.123 uV 13:13:05 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -254.967 uVs 13:13:05 Integrated Voltage Measurement... 13:13:14 Voltage Offset Correction: v_off_ini = 995.015 uV, v_off_fin = 997.519 uV 13:13:14 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 252.374 uVs 13:13:19 I1Y Average and RMS of Measurements VT = -249.595 +- 5.933 uVs I1Y = -49.917 +- 1.185 uTm 13:13:21 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 13:13:24 Integrated Voltage Measurement... 13:13:33 Voltage Offset Correction: v_off_ini = 984.938 uV, v_off_fin = 978.875 uV 13:13:33 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -76.115 uVs 13:13:33 Integrated Voltage Measurement... 13:13:42 Voltage Offset Correction: v_off_ini = 996.148 uV, v_off_fin = 1001.911 uV 13:13:42 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 88.138 uVs 13:13:42 Integrated Voltage Measurement... 13:13:51 Voltage Offset Correction: v_off_ini = 982.538 uV, v_off_fin = 983.124 uV 13:13:51 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -81.291 uVs 13:13:51 Integrated Voltage Measurement... 13:14:00 Voltage Offset Correction: v_off_ini = 1000.129 uV, v_off_fin = 1014.612 uV 13:14:00 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 86.803 uVs 13:14:00 Integrated Voltage Measurement... 13:14:09 Voltage Offset Correction: v_off_ini = 1009.627 uV, v_off_fin = 995.040 uV 13:14:09 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -86.605 uVs 13:14:09 Integrated Voltage Measurement... 13:14:18 Voltage Offset Correction: v_off_ini = 995.748 uV, v_off_fin = 982.342 uV 13:14:18 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 96.399 uVs 13:14:20 I2X Average and RMS of Measurements VT = -85.892 +- 4.044 uVs I2X = 7.902 +- 0.372 uTm^2 13:14:23 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 13:14:25 Integrated Voltage Measurement... 13:14:34 Voltage Offset Correction: v_off_ini = 999.388 uV, v_off_fin = 993.474 uV 13:14:34 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -306.139 uVs 13:14:35 Integrated Voltage Measurement... 13:14:43 Voltage Offset Correction: v_off_ini = 991.763 uV, v_off_fin = 1003.053 uV 13:14:44 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 346.522 uVs 13:14:44 Integrated Voltage Measurement... 13:14:53 Voltage Offset Correction: v_off_ini = 982.695 uV, v_off_fin = 1003.751 uV 13:14:53 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -326.051 uVs 13:14:53 Integrated Voltage Measurement... 13:15:02 Voltage Offset Correction: v_off_ini = 997.343 uV, v_off_fin = 1002.228 uV 13:15:02 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 353.618 uVs 13:15:02 Integrated Voltage Measurement... 13:15:11 Voltage Offset Correction: v_off_ini = 984.797 uV, v_off_fin = 989.124 uV 13:15:11 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -345.390 uVs 13:15:11 Integrated Voltage Measurement... 13:15:20 Voltage Offset Correction: v_off_ini = 1006.127 uV, v_off_fin = 986.469 uV 13:15:20 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 330.533 uVs 13:15:22 I2Y Average and RMS of Measurements VT = -334.709 +- 5.973 uVs I2Y = -30.793 +- 0.550 uTm^2 13:15:26 Wire Position: X0 = 0.0000 mm, Y0 = 0.5000 mm 13:15:30 Integrated Voltage Measurement... 13:15:39 Voltage Offset Correction: v_off_ini = 999.873 uV, v_off_fin = 993.135 uV 13:15:39 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.4999 mm dXY = -0.4999 mm VT = -54.299 uVs 13:15:39 Integrated Voltage Measurement... 13:15:48 Voltage Offset Correction: v_off_ini = 982.229 uV, v_off_fin = 984.947 uV 13:15:48 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 52.956 uVs 13:15:48 Integrated Voltage Measurement... 13:15:57 Voltage Offset Correction: v_off_ini = 999.248 uV, v_off_fin = 969.719 uV 13:15:57 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -41.186 uVs 13:15:57 Integrated Voltage Measurement... 13:16:06 Voltage Offset Correction: v_off_ini = 1000.820 uV, v_off_fin = 1002.317 uV 13:16:06 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 73.807 uVs 13:16:07 Integrated Voltage Measurement... 13:16:15 Voltage Offset Correction: v_off_ini = 1012.223 uV, v_off_fin = 999.968 uV 13:16:16 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -53.069 uVs 13:16:16 Integrated Voltage Measurement... 13:16:25 Voltage Offset Correction: v_off_ini = 1022.802 uV, v_off_fin = 1024.233 uV 13:16:25 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 42.155 uVs 13:16:29 I1X Average and RMS of Measurements VT = -52.912 +- 4.067 uVs I1X = 10.583 +- 0.813 uTm 13:16:31 Wire Position: X0 = 0.0000 mm, Y0 = 0.5000 mm 13:16:36 Integrated Voltage Measurement... 13:16:44 Voltage Offset Correction: v_off_ini = 1009.046 uV, v_off_fin = 1010.578 uV 13:16:45 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -229.131 uVs 13:16:45 Integrated Voltage Measurement... 13:16:53 Voltage Offset Correction: v_off_ini = 1036.293 uV, v_off_fin = 1032.180 uV 13:16:54 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 237.791 uVs 13:16:54 Integrated Voltage Measurement... 13:17:03 Voltage Offset Correction: v_off_ini = 1000.568 uV, v_off_fin = 1001.209 uV 13:17:03 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5002 mm dXY = 0.5001 mm VT = -256.477 uVs 13:17:03 Integrated Voltage Measurement... 13:17:12 Voltage Offset Correction: v_off_ini = 1017.452 uV, v_off_fin = 1017.309 uV 13:17:12 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5002 mm dXY = -0.5001 mm VT = 225.377 uVs 13:17:12 Integrated Voltage Measurement... 13:17:21 Voltage Offset Correction: v_off_ini = 1016.226 uV, v_off_fin = 1014.139 uV 13:17:21 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -249.317 uVs 13:17:21 Integrated Voltage Measurement... 13:17:30 Voltage Offset Correction: v_off_ini = 999.434 uV, v_off_fin = 1005.991 uV 13:17:30 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 233.566 uVs 13:17:34 I1Y Average and RMS of Measurements VT = -238.610 +- 3.647 uVs I1Y = -47.716 +- 0.727 uTm 13:17:36 Wire Position: X0 = 0.0000 mm, Y0 = 0.5000 mm 13:17:39 Integrated Voltage Measurement... 13:17:48 Voltage Offset Correction: v_off_ini = 1008.120 uV, v_off_fin = 1004.899 uV 13:17:48 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -71.196 uVs 13:17:48 Integrated Voltage Measurement... 13:17:57 Voltage Offset Correction: v_off_ini = 1002.941 uV, v_off_fin = 1008.997 uV 13:17:57 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 62.140 uVs 13:17:57 Integrated Voltage Measurement... 13:18:06 Voltage Offset Correction: v_off_ini = 999.458 uV, v_off_fin = 1008.695 uV 13:18:06 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -65.901 uVs 13:18:06 Integrated Voltage Measurement... 13:18:15 Voltage Offset Correction: v_off_ini = 1033.992 uV, v_off_fin = 1037.868 uV 13:18:15 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 72.177 uVs 13:18:15 Integrated Voltage Measurement... 13:18:24 Voltage Offset Correction: v_off_ini = 1037.195 uV, v_off_fin = 996.370 uV 13:18:24 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -60.673 uVs 13:18:24 Integrated Voltage Measurement... 13:18:33 Voltage Offset Correction: v_off_ini = 1008.793 uV, v_off_fin = 1012.682 uV 13:18:33 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 70.181 uVs 13:18:36 I2X Average and RMS of Measurements VT = -67.045 +- 1.499 uVs I2X = 6.168 +- 0.138 uTm^2 13:18:38 Wire Position: X0 = 0.0000 mm, Y0 = 0.5000 mm 13:18:41 Integrated Voltage Measurement... 13:18:50 Voltage Offset Correction: v_off_ini = 1003.890 uV, v_off_fin = 996.872 uV 13:18:50 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -320.467 uVs 13:18:50 Integrated Voltage Measurement... 13:18:59 Voltage Offset Correction: v_off_ini = 981.985 uV, v_off_fin = 995.469 uV 13:18:59 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 338.158 uVs 13:18:59 Integrated Voltage Measurement... 13:19:08 Voltage Offset Correction: v_off_ini = 988.336 uV, v_off_fin = 993.181 uV 13:19:08 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -339.347 uVs 13:19:08 Integrated Voltage Measurement... 13:19:17 Voltage Offset Correction: v_off_ini = 1012.115 uV, v_off_fin = 998.057 uV 13:19:17 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 330.581 uVs 13:19:17 Integrated Voltage Measurement... 13:19:26 Voltage Offset Correction: v_off_ini = 982.231 uV, v_off_fin = 1001.588 uV 13:19:26 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -310.676 uVs 13:19:26 Integrated Voltage Measurement... 13:19:35 Voltage Offset Correction: v_off_ini = 988.594 uV, v_off_fin = 1009.623 uV 13:19:35 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = 355.687 uVs 13:19:37 I2Y Average and RMS of Measurements VT = -332.486 +- 2.359 uVs I2Y = -30.589 +- 0.217 uTm^2 13:19:41 Wire Position: X0 = 0.0000 mm, Y0 = 1.0000 mm 13:19:46 Integrated Voltage Measurement... 13:19:55 Voltage Offset Correction: v_off_ini = 1004.957 uV, v_off_fin = 993.857 uV 13:19:55 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -67.188 uVs 13:19:55 Integrated Voltage Measurement... 13:20:04 Voltage Offset Correction: v_off_ini = 985.155 uV, v_off_fin = 998.504 uV 13:20:04 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 63.353 uVs 13:20:04 Integrated Voltage Measurement... 13:20:13 Voltage Offset Correction: v_off_ini = 964.704 uV, v_off_fin = 974.363 uV 13:20:13 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -63.440 uVs 13:20:13 Integrated Voltage Measurement... 13:20:22 Voltage Offset Correction: v_off_ini = 969.995 uV, v_off_fin = 968.182 uV 13:20:22 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 42.892 uVs 13:20:22 Integrated Voltage Measurement... 13:20:31 Voltage Offset Correction: v_off_ini = 980.942 uV, v_off_fin = 969.445 uV 13:20:31 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -67.955 uVs 13:20:31 Integrated Voltage Measurement... 13:20:40 Voltage Offset Correction: v_off_ini = 958.650 uV, v_off_fin = 973.322 uV 13:20:40 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 82.086 uVs 13:20:44 I1X Average and RMS of Measurements VT = -64.486 +- 8.939 uVs I1X = 12.897 +- 1.788 uTm 13:20:46 Wire Position: X0 = 0.0000 mm, Y0 = 1.0000 mm 13:20:51 Integrated Voltage Measurement... 13:21:00 Voltage Offset Correction: v_off_ini = 964.464 uV, v_off_fin = 956.514 uV 13:21:00 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.4999 mm dXY = 0.4999 mm VT = -242.647 uVs 13:21:00 Integrated Voltage Measurement... 13:21:09 Voltage Offset Correction: v_off_ini = 976.566 uV, v_off_fin = 1007.645 uV 13:21:09 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 241.136 uVs 13:21:09 Integrated Voltage Measurement... 13:21:18 Voltage Offset Correction: v_off_ini = 967.672 uV, v_off_fin = 973.573 uV 13:21:18 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -241.709 uVs 13:21:18 Integrated Voltage Measurement... 13:21:27 Voltage Offset Correction: v_off_ini = 983.106 uV, v_off_fin = 983.299 uV 13:21:27 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 236.480 uVs 13:21:27 Integrated Voltage Measurement... 13:21:36 Voltage Offset Correction: v_off_ini = 969.626 uV, v_off_fin = 938.979 uV 13:21:36 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5002 mm dXY = 0.5001 mm VT = -249.063 uVs 13:21:36 Integrated Voltage Measurement... 13:21:45 Voltage Offset Correction: v_off_ini = 960.053 uV, v_off_fin = 978.338 uV 13:21:45 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 250.367 uVs 13:21:49 I1Y Average and RMS of Measurements VT = -243.567 +- 4.495 uVs I1Y = -48.711 +- 0.896 uTm 13:21:51 Wire Position: X0 = 0.0000 mm, Y0 = 1.0000 mm 13:21:54 Integrated Voltage Measurement... 13:22:03 Voltage Offset Correction: v_off_ini = 972.101 uV, v_off_fin = 956.183 uV 13:22:03 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -88.456 uVs 13:22:03 Integrated Voltage Measurement... 13:22:12 Voltage Offset Correction: v_off_ini = 977.916 uV, v_off_fin = 976.267 uV 13:22:12 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 80.247 uVs 13:22:12 Integrated Voltage Measurement... 13:22:21 Voltage Offset Correction: v_off_ini = 983.524 uV, v_off_fin = 985.157 uV 13:22:21 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -83.640 uVs 13:22:21 Integrated Voltage Measurement... 13:22:30 Voltage Offset Correction: v_off_ini = 975.117 uV, v_off_fin = 990.100 uV 13:22:30 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 78.998 uVs 13:22:30 Integrated Voltage Measurement... 13:22:39 Voltage Offset Correction: v_off_ini = 967.661 uV, v_off_fin = 956.319 uV 13:22:39 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -56.363 uVs 13:22:39 Integrated Voltage Measurement... 13:22:48 Voltage Offset Correction: v_off_ini = 960.033 uV, v_off_fin = 978.211 uV 13:22:48 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 78.195 uVs 13:22:51 I2X Average and RMS of Measurements VT = -77.650 +- 7.437 uVs I2X = 7.144 +- 0.684 uTm^2 13:22:53 Wire Position: X0 = 0.0000 mm, Y0 = 1.0000 mm 13:22:56 Integrated Voltage Measurement... 13:23:05 Voltage Offset Correction: v_off_ini = 949.952 uV, v_off_fin = 954.326 uV 13:23:05 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = -331.648 uVs 13:23:05 Integrated Voltage Measurement... 13:23:14 Voltage Offset Correction: v_off_ini = 955.593 uV, v_off_fin = 963.712 uV 13:23:14 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = 330.823 uVs 13:23:14 Integrated Voltage Measurement... 13:23:23 Voltage Offset Correction: v_off_ini = 967.089 uV, v_off_fin = 971.663 uV 13:23:23 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -326.018 uVs 13:23:23 Integrated Voltage Measurement... 13:23:32 Voltage Offset Correction: v_off_ini = 977.125 uV, v_off_fin = 990.840 uV 13:23:32 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 312.522 uVs 13:23:32 Integrated Voltage Measurement... 13:23:41 Voltage Offset Correction: v_off_ini = 975.352 uV, v_off_fin = 970.735 uV 13:23:41 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -322.615 uVs 13:23:41 Integrated Voltage Measurement... 13:23:50 Voltage Offset Correction: v_off_ini = 959.194 uV, v_off_fin = 978.836 uV 13:23:50 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 350.220 uVs 13:23:52 I2Y Average and RMS of Measurements VT = -328.974 +- 7.181 uVs I2Y = -30.267 +- 0.661 uTm^2 13:24:08 Wire Position: X0 = -2.0000 mm, Y0 = 0.0000 mm 13:24:13 Integrated Voltage Measurement... 13:24:22 Voltage Offset Correction: v_off_ini = 975.879 uV, v_off_fin = 982.054 uV 13:24:22 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -80.004 uVs 13:24:22 Integrated Voltage Measurement... 13:24:31 Voltage Offset Correction: v_off_ini = 955.505 uV, v_off_fin = 969.017 uV 13:24:31 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 74.848 uVs 13:24:31 Integrated Voltage Measurement... 13:24:40 Voltage Offset Correction: v_off_ini = 972.731 uV, v_off_fin = 984.965 uV 13:24:40 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -70.711 uVs 13:24:40 Integrated Voltage Measurement... 13:24:49 Voltage Offset Correction: v_off_ini = 958.885 uV, v_off_fin = 953.344 uV 13:24:49 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 81.600 uVs 13:24:49 Integrated Voltage Measurement... 13:24:58 Voltage Offset Correction: v_off_ini = 962.493 uV, v_off_fin = 948.823 uV 13:24:58 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -61.012 uVs 13:24:58 Integrated Voltage Measurement... 13:25:07 Voltage Offset Correction: v_off_ini = 957.602 uV, v_off_fin = 943.647 uV 13:25:07 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 65.523 uVs 13:25:11 I1X Average and RMS of Measurements VT = -72.283 +- 6.396 uVs I1X = 14.457 +- 1.279 uTm 13:25:14 Wire Position: X0 = -2.0000 mm, Y0 = 0.0000 mm 13:25:18 Integrated Voltage Measurement... 13:25:27 Voltage Offset Correction: v_off_ini = 949.783 uV, v_off_fin = 954.527 uV 13:25:27 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -252.923 uVs 13:25:27 Integrated Voltage Measurement... 13:25:36 Voltage Offset Correction: v_off_ini = 967.886 uV, v_off_fin = 973.766 uV 13:25:36 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 242.911 uVs 13:25:36 Integrated Voltage Measurement... 13:25:45 Voltage Offset Correction: v_off_ini = 971.621 uV, v_off_fin = 950.989 uV 13:25:45 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -252.336 uVs 13:25:45 Integrated Voltage Measurement... 13:25:54 Voltage Offset Correction: v_off_ini = 963.045 uV, v_off_fin = 961.086 uV 13:25:54 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 255.911 uVs 13:25:54 Integrated Voltage Measurement... 13:26:03 Voltage Offset Correction: v_off_ini = 970.495 uV, v_off_fin = 950.488 uV 13:26:03 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -274.083 uVs 13:26:03 Integrated Voltage Measurement... 13:26:12 Voltage Offset Correction: v_off_ini = 962.351 uV, v_off_fin = 977.609 uV 13:26:12 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 261.307 uVs 13:26:16 I1Y Average and RMS of Measurements VT = -256.578 +- 8.259 uVs I1Y = -51.316 +- 1.652 uTm 13:26:21 Wire Position: X0 = -2.0000 mm, Y0 = 0.0000 mm 13:26:24 Integrated Voltage Measurement... 13:26:33 Voltage Offset Correction: v_off_ini = 953.932 uV, v_off_fin = 951.952 uV 13:26:33 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -91.947 uVs 13:26:33 Integrated Voltage Measurement... 13:26:42 Voltage Offset Correction: v_off_ini = 940.521 uV, v_off_fin = 965.137 uV 13:26:42 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 87.667 uVs 13:26:42 Integrated Voltage Measurement... 13:26:51 Voltage Offset Correction: v_off_ini = 961.921 uV, v_off_fin = 963.877 uV 13:26:51 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -65.129 uVs 13:26:51 Integrated Voltage Measurement... 13:27:00 Voltage Offset Correction: v_off_ini = 963.450 uV, v_off_fin = 950.393 uV 13:27:00 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 130.170 uVs 13:27:00 Integrated Voltage Measurement... 13:27:09 Voltage Offset Correction: v_off_ini = 976.912 uV, v_off_fin = 968.294 uV 13:27:09 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -91.419 uVs 13:27:09 Integrated Voltage Measurement... 13:27:18 Voltage Offset Correction: v_off_ini = 962.777 uV, v_off_fin = 971.991 uV 13:27:18 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 86.314 uVs 13:27:20 I2X Average and RMS of Measurements VT = -92.108 +- 3.937 uVs I2X = 8.474 +- 0.362 uTm^2 13:27:25 Wire Position: X0 = -2.0000 mm, Y0 = 0.0000 mm 13:27:28 Integrated Voltage Measurement... 13:27:37 Voltage Offset Correction: v_off_ini = 980.919 uV, v_off_fin = 981.644 uV 13:27:37 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -311.409 uVs 13:27:37 Integrated Voltage Measurement... 13:27:46 Voltage Offset Correction: v_off_ini = 990.010 uV, v_off_fin = 967.327 uV 13:27:46 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 362.269 uVs 13:27:46 Integrated Voltage Measurement... 13:27:55 Voltage Offset Correction: v_off_ini = 975.004 uV, v_off_fin = 951.567 uV 13:27:55 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -345.248 uVs 13:27:55 Integrated Voltage Measurement... 13:28:04 Voltage Offset Correction: v_off_ini = 963.192 uV, v_off_fin = 961.741 uV 13:28:04 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 361.803 uVs 13:28:04 Integrated Voltage Measurement... 13:28:13 Voltage Offset Correction: v_off_ini = 949.191 uV, v_off_fin = 971.731 uV 13:28:13 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -318.562 uVs 13:28:13 Integrated Voltage Measurement... 13:28:22 Voltage Offset Correction: v_off_ini = 966.981 uV, v_off_fin = 956.645 uV 13:28:22 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 364.820 uVs 13:28:25 I2Y Average and RMS of Measurements VT = -344.018 +- 7.008 uVs I2Y = -31.650 +- 0.645 uTm^2 13:28:29 Wire Position: X0 = -1.0000 mm, Y0 = 0.0000 mm 13:28:34 Integrated Voltage Measurement... 13:28:43 Voltage Offset Correction: v_off_ini = 957.548 uV, v_off_fin = 970.759 uV 13:28:43 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -63.847 uVs 13:28:43 Integrated Voltage Measurement... 13:28:52 Voltage Offset Correction: v_off_ini = 968.633 uV, v_off_fin = 962.297 uV 13:28:52 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 61.690 uVs 13:28:52 Integrated Voltage Measurement... 13:29:01 Voltage Offset Correction: v_off_ini = 965.508 uV, v_off_fin = 964.951 uV 13:29:01 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -58.659 uVs 13:29:01 Integrated Voltage Measurement... 13:29:10 Voltage Offset Correction: v_off_ini = 962.529 uV, v_off_fin = 964.321 uV 13:29:10 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 54.735 uVs 13:29:10 Integrated Voltage Measurement... 13:29:19 Voltage Offset Correction: v_off_ini = 956.769 uV, v_off_fin = 941.615 uV 13:29:19 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -69.422 uVs 13:29:19 Integrated Voltage Measurement... 13:29:28 Voltage Offset Correction: v_off_ini = 949.786 uV, v_off_fin = 951.543 uV 13:29:28 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 63.324 uVs 13:29:32 I1X Average and RMS of Measurements VT = -61.946 +- 3.993 uVs I1X = 12.389 +- 0.799 uTm 13:29:35 Wire Position: X0 = -1.0000 mm, Y0 = 0.0000 mm 13:29:39 Integrated Voltage Measurement... 13:29:48 Voltage Offset Correction: v_off_ini = 939.103 uV, v_off_fin = 918.640 uV 13:29:48 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -269.635 uVs 13:29:48 Integrated Voltage Measurement... 13:29:57 Voltage Offset Correction: v_off_ini = 925.683 uV, v_off_fin = 944.172 uV 13:29:57 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 241.176 uVs 13:29:57 Integrated Voltage Measurement... 13:30:06 Voltage Offset Correction: v_off_ini = 938.569 uV, v_off_fin = 929.960 uV 13:30:06 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -228.166 uVs 13:30:06 Integrated Voltage Measurement... 13:30:15 Voltage Offset Correction: v_off_ini = 935.444 uV, v_off_fin = 943.177 uV 13:30:15 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 262.878 uVs 13:30:15 Integrated Voltage Measurement... 13:30:24 Voltage Offset Correction: v_off_ini = 947.336 uV, v_off_fin = 928.226 uV 13:30:24 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -262.198 uVs 13:30:24 Integrated Voltage Measurement... 13:30:33 Voltage Offset Correction: v_off_ini = 954.034 uV, v_off_fin = 955.940 uV 13:30:33 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 255.802 uVs 13:30:37 I1Y Average and RMS of Measurements VT = -253.309 +- 5.699 uVs I1Y = -50.662 +- 1.140 uTm 13:30:42 Wire Position: X0 = -1.0000 mm, Y0 = 0.0000 mm 13:30:45 Integrated Voltage Measurement... 13:30:54 Voltage Offset Correction: v_off_ini = 927.939 uV, v_off_fin = 933.298 uV 13:30:54 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -70.888 uVs 13:30:54 Integrated Voltage Measurement... 13:31:03 Voltage Offset Correction: v_off_ini = 923.619 uV, v_off_fin = 929.582 uV 13:31:03 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 96.829 uVs 13:31:03 Integrated Voltage Measurement... 13:31:12 Voltage Offset Correction: v_off_ini = 949.298 uV, v_off_fin = 936.912 uV 13:31:12 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -70.482 uVs 13:31:12 Integrated Voltage Measurement... 13:31:21 Voltage Offset Correction: v_off_ini = 956.615 uV, v_off_fin = 961.781 uV 13:31:21 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 80.129 uVs 13:31:21 Integrated Voltage Measurement... 13:31:30 Voltage Offset Correction: v_off_ini = 947.102 uV, v_off_fin = 928.296 uV 13:31:30 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -90.398 uVs 13:31:30 Integrated Voltage Measurement... 13:31:39 Voltage Offset Correction: v_off_ini = 934.403 uV, v_off_fin = 952.420 uV 13:31:39 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 77.311 uVs 13:31:41 I2X Average and RMS of Measurements VT = -81.006 +- 4.031 uVs I2X = 7.453 +- 0.371 uTm^2 13:31:46 Wire Position: X0 = -1.0000 mm, Y0 = 0.0000 mm 13:31:49 Integrated Voltage Measurement... 13:31:58 Voltage Offset Correction: v_off_ini = 950.149 uV, v_off_fin = 943.576 uV 13:31:58 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -322.917 uVs 13:31:58 Integrated Voltage Measurement... 13:32:07 Voltage Offset Correction: v_off_ini = 936.576 uV, v_off_fin = 927.185 uV 13:32:07 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 303.892 uVs 13:32:07 Integrated Voltage Measurement... 13:32:16 Voltage Offset Correction: v_off_ini = 942.221 uV, v_off_fin = 954.817 uV 13:32:16 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -342.897 uVs 13:32:16 Integrated Voltage Measurement... 13:32:25 Voltage Offset Correction: v_off_ini = 932.426 uV, v_off_fin = 939.673 uV 13:32:25 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 328.437 uVs 13:32:25 Integrated Voltage Measurement... 13:32:34 Voltage Offset Correction: v_off_ini = 958.927 uV, v_off_fin = 958.887 uV 13:32:34 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -328.949 uVs 13:32:34 Integrated Voltage Measurement... 13:32:43 Voltage Offset Correction: v_off_ini = 978.752 uV, v_off_fin = 964.057 uV 13:32:43 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 309.046 uVs 13:32:46 I2Y Average and RMS of Measurements VT = -322.689 +- 9.456 uVs I2Y = -29.687 +- 0.870 uTm^2 13:32:50 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 13:32:55 Integrated Voltage Measurement... 13:33:04 Voltage Offset Correction: v_off_ini = 954.791 uV, v_off_fin = 940.035 uV 13:33:04 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -83.437 uVs 13:33:04 Integrated Voltage Measurement... 13:33:13 Voltage Offset Correction: v_off_ini = 953.907 uV, v_off_fin = 943.809 uV 13:33:13 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 61.103 uVs 13:33:13 Integrated Voltage Measurement... 13:33:22 Voltage Offset Correction: v_off_ini = 940.320 uV, v_off_fin = 940.136 uV 13:33:22 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -78.764 uVs 13:33:22 Integrated Voltage Measurement... 13:33:31 Voltage Offset Correction: v_off_ini = 944.328 uV, v_off_fin = 947.208 uV 13:33:31 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 53.861 uVs 13:33:31 Integrated Voltage Measurement... 13:33:40 Voltage Offset Correction: v_off_ini = 936.823 uV, v_off_fin = 949.753 uV 13:33:40 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -63.171 uVs 13:33:40 Integrated Voltage Measurement... 13:33:49 Voltage Offset Correction: v_off_ini = 947.349 uV, v_off_fin = 940.376 uV 13:33:49 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 39.915 uVs 13:33:53 I1X Average and RMS of Measurements VT = -63.375 +- 8.713 uVs I1X = 12.675 +- 1.743 uTm 13:33:56 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 13:34:00 Integrated Voltage Measurement... 13:34:09 Voltage Offset Correction: v_off_ini = 956.653 uV, v_off_fin = 953.278 uV 13:34:09 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = -252.950 uVs 13:34:09 Integrated Voltage Measurement... 13:34:18 Voltage Offset Correction: v_off_ini = 948.883 uV, v_off_fin = 951.920 uV 13:34:18 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 251.382 uVs 13:34:18 Integrated Voltage Measurement... 13:34:27 Voltage Offset Correction: v_off_ini = 950.951 uV, v_off_fin = 939.354 uV 13:34:27 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -244.403 uVs 13:34:27 Integrated Voltage Measurement... 13:34:36 Voltage Offset Correction: v_off_ini = 937.258 uV, v_off_fin = 952.972 uV 13:34:36 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 258.264 uVs 13:34:36 Integrated Voltage Measurement... 13:34:45 Voltage Offset Correction: v_off_ini = 944.027 uV, v_off_fin = 932.203 uV 13:34:45 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -250.254 uVs 13:34:45 Integrated Voltage Measurement... 13:34:54 Voltage Offset Correction: v_off_ini = 934.766 uV, v_off_fin = 964.013 uV 13:34:54 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 267.456 uVs 13:34:58 I1Y Average and RMS of Measurements VT = -254.118 +- 3.367 uVs I1Y = -50.824 +- 0.673 uTm 13:35:03 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 13:35:06 Integrated Voltage Measurement... 13:35:15 Voltage Offset Correction: v_off_ini = 944.918 uV, v_off_fin = 961.209 uV 13:35:15 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -88.243 uVs 13:35:15 Integrated Voltage Measurement... 13:35:24 Voltage Offset Correction: v_off_ini = 954.468 uV, v_off_fin = 954.070 uV 13:35:24 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 93.592 uVs 13:35:24 Integrated Voltage Measurement... 13:35:33 Voltage Offset Correction: v_off_ini = 940.924 uV, v_off_fin = 952.546 uV 13:35:33 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -64.138 uVs 13:35:33 Integrated Voltage Measurement... 13:35:42 Voltage Offset Correction: v_off_ini = 947.663 uV, v_off_fin = 951.026 uV 13:35:42 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 89.772 uVs 13:35:42 Integrated Voltage Measurement... 13:35:51 Voltage Offset Correction: v_off_ini = 948.235 uV, v_off_fin = 942.996 uV 13:35:51 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -85.309 uVs 13:35:51 Integrated Voltage Measurement... 13:36:00 Voltage Offset Correction: v_off_ini = 948.504 uV, v_off_fin = 953.640 uV 13:36:00 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 62.311 uVs 13:36:02 I2X Average and RMS of Measurements VT = -80.561 +- 7.435 uVs I2X = 7.412 +- 0.684 uTm^2 13:36:07 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 13:36:10 Integrated Voltage Measurement... 13:36:19 Voltage Offset Correction: v_off_ini = 950.932 uV, v_off_fin = 939.956 uV 13:36:19 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = -317.490 uVs 13:36:19 Integrated Voltage Measurement... 13:36:28 Voltage Offset Correction: v_off_ini = 948.941 uV, v_off_fin = 935.461 uV 13:36:28 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 309.059 uVs 13:36:28 Integrated Voltage Measurement... 13:36:37 Voltage Offset Correction: v_off_ini = 935.529 uV, v_off_fin = 937.222 uV 13:36:37 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -331.760 uVs 13:36:37 Integrated Voltage Measurement... 13:36:46 Voltage Offset Correction: v_off_ini = 941.091 uV, v_off_fin = 919.995 uV 13:36:46 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 297.961 uVs 13:36:46 Integrated Voltage Measurement... 13:36:55 Voltage Offset Correction: v_off_ini = 933.760 uV, v_off_fin = 937.409 uV 13:36:55 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -340.078 uVs 13:36:55 Integrated Voltage Measurement... 13:37:04 Voltage Offset Correction: v_off_ini = 945.275 uV, v_off_fin = 942.271 uV 13:37:04 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 320.613 uVs 13:37:07 I2Y Average and RMS of Measurements VT = -319.493 +- 7.701 uVs I2Y = -29.394 +- 0.708 uTm^2 13:37:11 Wire Position: X0 = 1.0000 mm, Y0 = 0.0000 mm 13:37:16 Integrated Voltage Measurement... 13:37:25 Voltage Offset Correction: v_off_ini = 946.646 uV, v_off_fin = 947.085 uV 13:37:25 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -58.853 uVs 13:37:25 Integrated Voltage Measurement... 13:37:34 Voltage Offset Correction: v_off_ini = 940.026 uV, v_off_fin = 931.368 uV 13:37:34 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 78.214 uVs 13:37:34 Integrated Voltage Measurement... 13:37:43 Voltage Offset Correction: v_off_ini = 944.421 uV, v_off_fin = 918.474 uV 13:37:43 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -56.789 uVs 13:37:43 Integrated Voltage Measurement... 13:37:52 Voltage Offset Correction: v_off_ini = 936.661 uV, v_off_fin = 926.426 uV 13:37:52 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 78.523 uVs 13:37:52 Integrated Voltage Measurement... 13:38:01 Voltage Offset Correction: v_off_ini = 927.536 uV, v_off_fin = 931.736 uV 13:38:01 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -72.259 uVs 13:38:01 Integrated Voltage Measurement... 13:38:10 Voltage Offset Correction: v_off_ini = 937.782 uV, v_off_fin = 955.074 uV 13:38:10 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 71.175 uVs 13:38:14 I1X Average and RMS of Measurements VT = -69.302 +- 1.745 uVs I1X = 13.861 +- 0.349 uTm 13:38:16 Wire Position: X0 = 1.0000 mm, Y0 = 0.0000 mm 13:38:21 Integrated Voltage Measurement... 13:38:30 Voltage Offset Correction: v_off_ini = 941.043 uV, v_off_fin = 932.114 uV 13:38:30 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -255.363 uVs 13:38:30 Integrated Voltage Measurement... 13:38:39 Voltage Offset Correction: v_off_ini = 931.004 uV, v_off_fin = 952.004 uV 13:38:39 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 261.965 uVs 13:38:39 Integrated Voltage Measurement... 13:38:48 Voltage Offset Correction: v_off_ini = 926.261 uV, v_off_fin = 908.303 uV 13:38:48 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -251.689 uVs 13:38:48 Integrated Voltage Measurement... 13:38:57 Voltage Offset Correction: v_off_ini = 938.471 uV, v_off_fin = 953.102 uV 13:38:57 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 237.723 uVs 13:38:57 Integrated Voltage Measurement... 13:39:06 Voltage Offset Correction: v_off_ini = 957.964 uV, v_off_fin = 952.818 uV 13:39:06 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -253.039 uVs 13:39:06 Integrated Voltage Measurement... 13:39:15 Voltage Offset Correction: v_off_ini = 977.096 uV, v_off_fin = 948.246 uV 13:39:15 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 223.628 uVs 13:39:19 I1Y Average and RMS of Measurements VT = -247.235 +- 8.490 uVs I1Y = -49.447 +- 1.698 uTm 13:39:24 Wire Position: X0 = 1.0000 mm, Y0 = 0.0000 mm 13:39:27 Integrated Voltage Measurement... 13:39:36 Voltage Offset Correction: v_off_ini = 958.712 uV, v_off_fin = 946.381 uV 13:39:36 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -95.648 uVs 13:39:36 Integrated Voltage Measurement... 13:39:45 Voltage Offset Correction: v_off_ini = 959.068 uV, v_off_fin = 977.447 uV 13:39:45 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 72.824 uVs 13:39:45 Integrated Voltage Measurement... 13:39:54 Voltage Offset Correction: v_off_ini = 973.861 uV, v_off_fin = 950.015 uV 13:39:54 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -69.030 uVs 13:39:54 Integrated Voltage Measurement... 13:40:03 Voltage Offset Correction: v_off_ini = 950.096 uV, v_off_fin = 958.740 uV 13:40:03 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 83.440 uVs 13:40:03 Integrated Voltage Measurement... 13:40:12 Voltage Offset Correction: v_off_ini = 962.391 uV, v_off_fin = 961.014 uV 13:40:12 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -95.265 uVs 13:40:12 Integrated Voltage Measurement... 13:40:21 Voltage Offset Correction: v_off_ini = 962.965 uV, v_off_fin = 971.548 uV 13:40:21 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 62.971 uVs 13:40:23 I2X Average and RMS of Measurements VT = -79.863 +- 3.309 uVs I2X = 7.348 +- 0.304 uTm^2 13:40:28 Wire Position: X0 = 1.0000 mm, Y0 = 0.0000 mm 13:40:31 Integrated Voltage Measurement... 13:40:40 Voltage Offset Correction: v_off_ini = 961.606 uV, v_off_fin = 956.275 uV 13:40:40 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -310.574 uVs 13:40:40 Integrated Voltage Measurement... 13:40:49 Voltage Offset Correction: v_off_ini = 948.428 uV, v_off_fin = 965.694 uV 13:40:49 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 334.034 uVs 13:40:49 Integrated Voltage Measurement... 13:40:58 Voltage Offset Correction: v_off_ini = 967.767 uV, v_off_fin = 982.986 uV 13:40:58 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -326.981 uVs 13:40:58 Integrated Voltage Measurement... 13:41:07 Voltage Offset Correction: v_off_ini = 961.196 uV, v_off_fin = 944.811 uV 13:41:07 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 338.866 uVs 13:41:07 Integrated Voltage Measurement... 13:41:16 Voltage Offset Correction: v_off_ini = 964.328 uV, v_off_fin = 972.768 uV 13:41:16 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -318.681 uVs 13:41:16 Integrated Voltage Measurement... 13:41:25 Voltage Offset Correction: v_off_ini = 957.201 uV, v_off_fin = 965.567 uV 13:41:25 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 326.722 uVs 13:41:28 I2Y Average and RMS of Measurements VT = -325.976 +- 4.915 uVs I2Y = -29.990 +- 0.452 uTm^2 13:41:32 Wire Position: X0 = 2.0000 mm, Y0 = 0.0000 mm 13:41:37 Integrated Voltage Measurement... 13:41:46 Voltage Offset Correction: v_off_ini = 972.850 uV, v_off_fin = 956.589 uV 13:41:46 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -68.932 uVs 13:41:46 Integrated Voltage Measurement... 13:41:55 Voltage Offset Correction: v_off_ini = 960.371 uV, v_off_fin = 953.891 uV 13:41:55 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 75.666 uVs 13:41:55 Integrated Voltage Measurement... 13:42:04 Voltage Offset Correction: v_off_ini = 949.973 uV, v_off_fin = 955.145 uV 13:42:04 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -81.587 uVs 13:42:04 Integrated Voltage Measurement... 13:42:13 Voltage Offset Correction: v_off_ini = 964.244 uV, v_off_fin = 960.352 uV 13:42:13 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 59.508 uVs 13:42:13 Integrated Voltage Measurement... 13:42:22 Voltage Offset Correction: v_off_ini = 955.059 uV, v_off_fin = 955.663 uV 13:42:22 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -57.768 uVs 13:42:22 Integrated Voltage Measurement... 13:42:31 Voltage Offset Correction: v_off_ini = 949.735 uV, v_off_fin = 966.118 uV 13:42:31 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 71.937 uVs 13:42:35 I1X Average and RMS of Measurements VT = -69.233 +- 3.179 uVs I1X = 13.847 +- 0.636 uTm 13:42:37 Wire Position: X0 = 2.0000 mm, Y0 = 0.0000 mm 13:42:42 Integrated Voltage Measurement... 13:42:51 Voltage Offset Correction: v_off_ini = 943.768 uV, v_off_fin = 946.463 uV 13:42:51 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -245.766 uVs 13:42:51 Integrated Voltage Measurement... 13:43:00 Voltage Offset Correction: v_off_ini = 966.468 uV, v_off_fin = 978.247 uV 13:43:00 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.5000 mm dXY = -0.4999 mm VT = 234.617 uVs 13:43:00 Integrated Voltage Measurement... 13:43:09 Voltage Offset Correction: v_off_ini = 980.747 uV, v_off_fin = 973.061 uV 13:43:09 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -252.735 uVs 13:43:09 Integrated Voltage Measurement... 13:43:18 Voltage Offset Correction: v_off_ini = 960.633 uV, v_off_fin = 984.967 uV 13:43:18 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 242.838 uVs 13:43:18 Integrated Voltage Measurement... 13:43:27 Voltage Offset Correction: v_off_ini = 963.671 uV, v_off_fin = 961.174 uV 13:43:27 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -258.533 uVs 13:43:27 Integrated Voltage Measurement... 13:43:36 Voltage Offset Correction: v_off_ini = 970.824 uV, v_off_fin = 988.586 uV 13:43:36 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 229.398 uVs 13:43:40 I1Y Average and RMS of Measurements VT = -243.981 +- 3.101 uVs I1Y = -48.797 +- 0.619 uTm 13:43:45 Wire Position: X0 = 2.0000 mm, Y0 = 0.0000 mm 13:43:48 Integrated Voltage Measurement... 13:43:57 Voltage Offset Correction: v_off_ini = 971.090 uV, v_off_fin = 966.928 uV 13:43:57 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -72.669 uVs 13:43:57 Integrated Voltage Measurement... 13:44:06 Voltage Offset Correction: v_off_ini = 970.622 uV, v_off_fin = 983.954 uV 13:44:06 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 94.305 uVs 13:44:06 Integrated Voltage Measurement... 13:44:15 Voltage Offset Correction: v_off_ini = 966.483 uV, v_off_fin = 948.821 uV 13:44:15 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -82.378 uVs 13:44:15 Integrated Voltage Measurement... 13:44:24 Voltage Offset Correction: v_off_ini = 951.718 uV, v_off_fin = 968.409 uV 13:44:24 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 97.126 uVs 13:44:24 Integrated Voltage Measurement... 13:44:33 Voltage Offset Correction: v_off_ini = 937.615 uV, v_off_fin = 931.384 uV 13:44:33 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -82.148 uVs 13:44:33 Integrated Voltage Measurement... 13:44:42 Voltage Offset Correction: v_off_ini = 953.741 uV, v_off_fin = 947.049 uV 13:44:42 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 88.902 uVs 13:44:44 I2X Average and RMS of Measurements VT = -86.255 +- 2.609 uVs I2X = 7.936 +- 0.240 uTm^2 13:44:49 Wire Position: X0 = 2.0000 mm, Y0 = 0.0000 mm 13:44:52 Integrated Voltage Measurement... 13:45:01 Voltage Offset Correction: v_off_ini = 932.342 uV, v_off_fin = 944.962 uV 13:45:01 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -304.326 uVs 13:45:01 Integrated Voltage Measurement... 13:45:10 Voltage Offset Correction: v_off_ini = 938.460 uV, v_off_fin = 933.554 uV 13:45:10 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 304.065 uVs 13:45:10 Integrated Voltage Measurement... 13:45:19 Voltage Offset Correction: v_off_ini = 943.228 uV, v_off_fin = 953.768 uV 13:45:19 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -325.546 uVs 13:45:19 Integrated Voltage Measurement... 13:45:28 Voltage Offset Correction: v_off_ini = 962.853 uV, v_off_fin = 949.995 uV 13:45:28 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 343.190 uVs 13:45:28 Integrated Voltage Measurement... 13:45:37 Voltage Offset Correction: v_off_ini = 937.814 uV, v_off_fin = 940.480 uV 13:45:37 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -312.232 uVs 13:45:37 Integrated Voltage Measurement... 13:45:46 Voltage Offset Correction: v_off_ini = 919.463 uV, v_off_fin = 946.921 uV 13:45:46 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 352.561 uVs 13:45:49 I2Y Average and RMS of Measurements VT = -323.653 +- 13.782 uVs I2Y = -29.776 +- 1.268 uTm^2 13:45:56 Wire Position: X0 = 0.0000 mm, Y0 = -1.0000 mm 13:46:01 Integrated Voltage Measurement... 13:46:10 Voltage Offset Correction: v_off_ini = 956.697 uV, v_off_fin = 953.389 uV 13:46:10 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -67.563 uVs 13:46:10 Integrated Voltage Measurement... 13:46:19 Voltage Offset Correction: v_off_ini = 944.252 uV, v_off_fin = 942.180 uV 13:46:19 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 94.883 uVs 13:46:19 Integrated Voltage Measurement... 13:46:28 Voltage Offset Correction: v_off_ini = 938.546 uV, v_off_fin = 955.547 uV 13:46:28 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -102.038 uVs 13:46:28 Integrated Voltage Measurement... 13:46:37 Voltage Offset Correction: v_off_ini = 938.521 uV, v_off_fin = 932.127 uV 13:46:37 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 74.900 uVs 13:46:37 Integrated Voltage Measurement... 13:46:46 Voltage Offset Correction: v_off_ini = 934.224 uV, v_off_fin = 929.342 uV 13:46:46 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -76.298 uVs 13:46:46 Integrated Voltage Measurement... 13:46:55 Voltage Offset Correction: v_off_ini = 940.365 uV, v_off_fin = 951.598 uV 13:46:55 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 71.502 uVs 13:46:59 I1X Average and RMS of Measurements VT = -81.197 +- 5.948 uVs I1X = 16.240 +- 1.190 uTm 13:47:01 Wire Position: X0 = 0.0000 mm, Y0 = -1.0000 mm 13:47:06 Integrated Voltage Measurement... 13:47:15 Voltage Offset Correction: v_off_ini = 937.001 uV, v_off_fin = 928.480 uV 13:47:15 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -253.943 uVs 13:47:15 Integrated Voltage Measurement... 13:47:24 Voltage Offset Correction: v_off_ini = 919.775 uV, v_off_fin = 944.711 uV 13:47:24 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 258.549 uVs 13:47:24 Integrated Voltage Measurement... 13:47:33 Voltage Offset Correction: v_off_ini = 934.945 uV, v_off_fin = 926.690 uV 13:47:33 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -260.083 uVs 13:47:33 Integrated Voltage Measurement... 13:47:42 Voltage Offset Correction: v_off_ini = 929.935 uV, v_off_fin = 937.266 uV 13:47:42 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 253.997 uVs 13:47:42 Integrated Voltage Measurement... 13:47:51 Voltage Offset Correction: v_off_ini = 937.303 uV, v_off_fin = 916.989 uV 13:47:51 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -261.963 uVs 13:47:51 Integrated Voltage Measurement... 13:48:00 Voltage Offset Correction: v_off_ini = 932.487 uV, v_off_fin = 937.443 uV 13:48:00 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 263.755 uVs 13:48:04 I1Y Average and RMS of Measurements VT = -258.715 +- 2.948 uVs I1Y = -51.743 +- 0.590 uTm 13:48:06 Wire Position: X0 = 0.0000 mm, Y0 = -1.0000 mm 13:48:09 Integrated Voltage Measurement... 13:48:18 Voltage Offset Correction: v_off_ini = 932.706 uV, v_off_fin = 909.253 uV 13:48:18 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -87.409 uVs 13:48:18 Integrated Voltage Measurement... 13:48:27 Voltage Offset Correction: v_off_ini = 925.552 uV, v_off_fin = 929.670 uV 13:48:27 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 103.545 uVs 13:48:27 Integrated Voltage Measurement... 13:48:36 Voltage Offset Correction: v_off_ini = 896.947 uV, v_off_fin = 898.343 uV 13:48:36 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -109.727 uVs 13:48:36 Integrated Voltage Measurement... 13:48:45 Voltage Offset Correction: v_off_ini = 907.563 uV, v_off_fin = 924.790 uV 13:48:45 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 107.349 uVs 13:48:45 Integrated Voltage Measurement... 13:48:54 Voltage Offset Correction: v_off_ini = 924.044 uV, v_off_fin = 928.936 uV 13:48:54 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -95.282 uVs 13:48:54 Integrated Voltage Measurement... 13:49:03 Voltage Offset Correction: v_off_ini = 920.587 uV, v_off_fin = 936.275 uV 13:49:03 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 86.344 uVs 13:49:06 I2X Average and RMS of Measurements VT = -98.276 +- 7.502 uVs I2X = 9.042 +- 0.690 uTm^2 13:49:08 Wire Position: X0 = 0.0000 mm, Y0 = -1.0000 mm 13:49:11 Integrated Voltage Measurement... 13:49:20 Voltage Offset Correction: v_off_ini = 941.972 uV, v_off_fin = 945.714 uV 13:49:20 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -351.005 uVs 13:49:20 Integrated Voltage Measurement... 13:49:29 Voltage Offset Correction: v_off_ini = 929.859 uV, v_off_fin = 929.863 uV 13:49:29 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 346.238 uVs 13:49:29 Integrated Voltage Measurement... 13:49:38 Voltage Offset Correction: v_off_ini = 926.754 uV, v_off_fin = 950.320 uV 13:49:38 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -315.814 uVs 13:49:38 Integrated Voltage Measurement... 13:49:47 Voltage Offset Correction: v_off_ini = 923.575 uV, v_off_fin = 927.532 uV 13:49:47 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 357.984 uVs 13:49:47 Integrated Voltage Measurement... 13:49:56 Voltage Offset Correction: v_off_ini = 931.155 uV, v_off_fin = 917.553 uV 13:49:56 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -344.820 uVs 13:49:56 Integrated Voltage Measurement... 13:50:05 Voltage Offset Correction: v_off_ini = 919.096 uV, v_off_fin = 909.313 uV 13:50:05 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 342.003 uVs 13:50:07 I2Y Average and RMS of Measurements VT = -342.977 +- 4.795 uVs I2Y = -31.554 +- 0.441 uTm^2 13:50:12 Wire Position: X0 = 0.0000 mm, Y0 = -0.5000 mm 13:50:16 Integrated Voltage Measurement... 13:50:25 Voltage Offset Correction: v_off_ini = 913.602 uV, v_off_fin = 908.373 uV 13:50:25 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -99.701 uVs 13:50:25 Integrated Voltage Measurement... 13:50:34 Voltage Offset Correction: v_off_ini = 901.042 uV, v_off_fin = 915.212 uV 13:50:34 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 63.628 uVs 13:50:34 Integrated Voltage Measurement... 13:50:43 Voltage Offset Correction: v_off_ini = 903.480 uV, v_off_fin = 917.003 uV 13:50:43 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -71.449 uVs 13:50:43 Integrated Voltage Measurement... 13:50:52 Voltage Offset Correction: v_off_ini = 908.098 uV, v_off_fin = 910.437 uV 13:50:52 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 73.486 uVs 13:50:52 Integrated Voltage Measurement... 13:51:01 Voltage Offset Correction: v_off_ini = 892.998 uV, v_off_fin = 909.415 uV 13:51:01 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -74.421 uVs 13:51:01 Integrated Voltage Measurement... 13:51:10 Voltage Offset Correction: v_off_ini = 897.411 uV, v_off_fin = 900.160 uV 13:51:10 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 88.163 uVs 13:51:14 I1X Average and RMS of Measurements VT = -78.475 +- 4.251 uVs I1X = 15.695 +- 0.850 uTm 13:51:17 Wire Position: X0 = 0.0000 mm, Y0 = -0.5000 mm 13:51:21 Integrated Voltage Measurement... 13:51:30 Voltage Offset Correction: v_off_ini = 904.593 uV, v_off_fin = 909.432 uV 13:51:30 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -253.121 uVs 13:51:30 Integrated Voltage Measurement... 13:51:39 Voltage Offset Correction: v_off_ini = 922.472 uV, v_off_fin = 940.722 uV 13:51:39 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 269.983 uVs 13:51:39 Integrated Voltage Measurement... 13:51:48 Voltage Offset Correction: v_off_ini = 910.236 uV, v_off_fin = 901.890 uV 13:51:48 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -249.139 uVs 13:51:48 Integrated Voltage Measurement... 13:51:57 Voltage Offset Correction: v_off_ini = 922.404 uV, v_off_fin = 922.901 uV 13:51:57 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 224.124 uVs 13:51:57 Integrated Voltage Measurement... 13:52:06 Voltage Offset Correction: v_off_ini = 907.512 uV, v_off_fin = 910.508 uV 13:52:06 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -237.048 uVs 13:52:06 Integrated Voltage Measurement... 13:52:15 Voltage Offset Correction: v_off_ini = 917.161 uV, v_off_fin = 904.006 uV 13:52:15 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 252.498 uVs 13:52:19 I1Y Average and RMS of Measurements VT = -247.652 +- 10.375 uVs I1Y = -49.530 +- 2.075 uTm 13:52:22 Wire Position: X0 = 0.0000 mm, Y0 = -0.5000 mm 13:52:24 Integrated Voltage Measurement... 13:52:33 Voltage Offset Correction: v_off_ini = 890.808 uV, v_off_fin = 893.693 uV 13:52:33 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -74.726 uVs 13:52:33 Integrated Voltage Measurement... 13:52:42 Voltage Offset Correction: v_off_ini = 903.450 uV, v_off_fin = 912.935 uV 13:52:42 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 88.524 uVs 13:52:42 Integrated Voltage Measurement... 13:52:51 Voltage Offset Correction: v_off_ini = 915.947 uV, v_off_fin = 897.976 uV 13:52:51 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -107.389 uVs 13:52:51 Integrated Voltage Measurement... 13:53:00 Voltage Offset Correction: v_off_ini = 895.299 uV, v_off_fin = 891.076 uV 13:53:00 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 97.908 uVs 13:53:00 Integrated Voltage Measurement... 13:53:09 Voltage Offset Correction: v_off_ini = 890.640 uV, v_off_fin = 893.977 uV 13:53:09 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -89.330 uVs 13:53:10 Integrated Voltage Measurement... 13:53:18 Voltage Offset Correction: v_off_ini = 894.807 uV, v_off_fin = 916.174 uV 13:53:19 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 95.669 uVs 13:53:21 I2X Average and RMS of Measurements VT = -92.258 +- 8.585 uVs I2X = 8.488 +- 0.790 uTm^2 13:53:23 Wire Position: X0 = 0.0000 mm, Y0 = -0.5000 mm 13:53:26 Integrated Voltage Measurement... 13:53:35 Voltage Offset Correction: v_off_ini = 891.327 uV, v_off_fin = 901.100 uV 13:53:35 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -317.760 uVs 13:53:35 Integrated Voltage Measurement... 13:53:44 Voltage Offset Correction: v_off_ini = 908.992 uV, v_off_fin = 911.431 uV 13:53:44 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 331.406 uVs 13:53:44 Integrated Voltage Measurement... 13:53:53 Voltage Offset Correction: v_off_ini = 903.451 uV, v_off_fin = 882.480 uV 13:53:53 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -331.282 uVs 13:53:53 Integrated Voltage Measurement... 13:54:02 Voltage Offset Correction: v_off_ini = 908.884 uV, v_off_fin = 895.909 uV 13:54:02 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 336.485 uVs 13:54:02 Integrated Voltage Measurement... 13:54:11 Voltage Offset Correction: v_off_ini = 897.428 uV, v_off_fin = 902.158 uV 13:54:11 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -328.581 uVs 13:54:11 Integrated Voltage Measurement... 13:54:20 Voltage Offset Correction: v_off_ini = 916.005 uV, v_off_fin = 907.440 uV 13:54:20 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 320.216 uVs 13:54:23 I2Y Average and RMS of Measurements VT = -327.622 +- 4.428 uVs I2Y = -30.141 +- 0.407 uTm^2 13:54:27 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 13:54:31 Integrated Voltage Measurement... 13:54:40 Voltage Offset Correction: v_off_ini = 927.285 uV, v_off_fin = 919.094 uV 13:54:40 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -70.203 uVs 13:54:40 Integrated Voltage Measurement... 13:54:49 Voltage Offset Correction: v_off_ini = 909.110 uV, v_off_fin = 911.226 uV 13:54:49 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 71.354 uVs 13:54:49 Integrated Voltage Measurement... 13:54:58 Voltage Offset Correction: v_off_ini = 888.458 uV, v_off_fin = 890.801 uV 13:54:58 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -71.690 uVs 13:54:58 Integrated Voltage Measurement... 13:55:07 Voltage Offset Correction: v_off_ini = 885.477 uV, v_off_fin = 916.445 uV 13:55:07 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 54.573 uVs 13:55:07 Integrated Voltage Measurement... 13:55:16 Voltage Offset Correction: v_off_ini = 898.803 uV, v_off_fin = 908.268 uV 13:55:16 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -52.487 uVs 13:55:16 Integrated Voltage Measurement... 13:55:25 Voltage Offset Correction: v_off_ini = 902.708 uV, v_off_fin = 921.578 uV 13:55:25 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 69.208 uVs 13:55:29 I1X Average and RMS of Measurements VT = -64.919 +- 4.247 uVs I1X = 12.984 +- 0.850 uTm 13:55:32 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 13:55:36 Integrated Voltage Measurement... 13:55:45 Voltage Offset Correction: v_off_ini = 877.235 uV, v_off_fin = 902.496 uV 13:55:45 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -234.845 uVs 13:55:45 Integrated Voltage Measurement... 13:55:54 Voltage Offset Correction: v_off_ini = 900.131 uV, v_off_fin = 919.168 uV 13:55:54 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 248.580 uVs 13:55:54 Integrated Voltage Measurement... 13:56:03 Voltage Offset Correction: v_off_ini = 913.881 uV, v_off_fin = 882.067 uV 13:56:03 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -268.731 uVs 13:56:03 Integrated Voltage Measurement... 13:56:12 Voltage Offset Correction: v_off_ini = 879.759 uV, v_off_fin = 908.273 uV 13:56:12 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 246.321 uVs 13:56:12 Integrated Voltage Measurement... 13:56:21 Voltage Offset Correction: v_off_ini = 878.342 uV, v_off_fin = 859.935 uV 13:56:21 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -261.108 uVs 13:56:21 Integrated Voltage Measurement... 13:56:30 Voltage Offset Correction: v_off_ini = 865.266 uV, v_off_fin = 878.348 uV 13:56:30 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 265.250 uVs 13:56:34 I1Y Average and RMS of Measurements VT = -254.139 +- 9.085 uVs I1Y = -50.828 +- 1.817 uTm 13:56:37 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 13:56:39 Integrated Voltage Measurement... 13:56:48 Voltage Offset Correction: v_off_ini = 881.667 uV, v_off_fin = 883.088 uV 13:56:48 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -94.944 uVs 13:56:48 Integrated Voltage Measurement... 13:56:57 Voltage Offset Correction: v_off_ini = 889.880 uV, v_off_fin = 904.371 uV 13:56:57 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 80.307 uVs 13:56:58 Integrated Voltage Measurement... 13:57:07 Voltage Offset Correction: v_off_ini = 898.002 uV, v_off_fin = 884.270 uV 13:57:07 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -73.059 uVs 13:57:07 Integrated Voltage Measurement... 13:57:16 Voltage Offset Correction: v_off_ini = 890.586 uV, v_off_fin = 893.731 uV 13:57:16 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 88.776 uVs 13:57:16 Integrated Voltage Measurement... 13:57:25 Voltage Offset Correction: v_off_ini = 906.521 uV, v_off_fin = 902.671 uV 13:57:25 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -67.963 uVs 13:57:25 Integrated Voltage Measurement... 13:57:34 Voltage Offset Correction: v_off_ini = 901.896 uV, v_off_fin = 890.547 uV 13:57:34 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 74.175 uVs 13:57:36 I2X Average and RMS of Measurements VT = -79.871 +- 6.799 uVs I2X = 7.348 +- 0.626 uTm^2 13:57:39 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 13:57:41 Integrated Voltage Measurement... 13:57:50 Voltage Offset Correction: v_off_ini = 908.220 uV, v_off_fin = 904.833 uV 13:57:50 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -328.683 uVs 13:57:50 Integrated Voltage Measurement... 13:57:59 Voltage Offset Correction: v_off_ini = 902.765 uV, v_off_fin = 893.990 uV 13:57:59 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 308.207 uVs 13:57:59 Integrated Voltage Measurement... 13:58:08 Voltage Offset Correction: v_off_ini = 901.104 uV, v_off_fin = 905.023 uV 13:58:08 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -334.669 uVs 13:58:08 Integrated Voltage Measurement... 13:58:17 Voltage Offset Correction: v_off_ini = 894.054 uV, v_off_fin = 886.086 uV 13:58:17 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 337.632 uVs 13:58:17 Integrated Voltage Measurement... 13:58:26 Voltage Offset Correction: v_off_ini = 880.577 uV, v_off_fin = 893.485 uV 13:58:26 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -298.143 uVs 13:58:26 Integrated Voltage Measurement... 13:58:35 Voltage Offset Correction: v_off_ini = 909.627 uV, v_off_fin = 901.897 uV 13:58:35 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 305.802 uVs 13:58:38 I2Y Average and RMS of Measurements VT = -318.856 +- 13.956 uVs I2Y = -29.335 +- 1.284 uTm^2 13:58:42 Wire Position: X0 = 0.0000 mm, Y0 = 0.5000 mm 13:58:46 Integrated Voltage Measurement... 13:58:55 Voltage Offset Correction: v_off_ini = 911.794 uV, v_off_fin = 912.400 uV 13:58:55 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -78.941 uVs 13:58:55 Integrated Voltage Measurement... 13:59:04 Voltage Offset Correction: v_off_ini = 898.235 uV, v_off_fin = 901.596 uV 13:59:04 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 32.082 uVs 13:59:04 Integrated Voltage Measurement... 13:59:13 Voltage Offset Correction: v_off_ini = 891.067 uV, v_off_fin = 880.343 uV 13:59:13 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -45.249 uVs 13:59:13 Integrated Voltage Measurement... 13:59:22 Voltage Offset Correction: v_off_ini = 897.072 uV, v_off_fin = 891.476 uV 13:59:22 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 60.497 uVs 13:59:22 Integrated Voltage Measurement... 13:59:31 Voltage Offset Correction: v_off_ini = 885.950 uV, v_off_fin = 885.448 uV 13:59:31 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -47.242 uVs 13:59:31 Integrated Voltage Measurement... 13:59:40 Voltage Offset Correction: v_off_ini = 876.923 uV, v_off_fin = 877.827 uV 13:59:40 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 72.680 uVs 13:59:44 I1X Average and RMS of Measurements VT = -56.115 +- 2.925 uVs I1X = 11.223 +- 0.585 uTm 13:59:47 Wire Position: X0 = 0.0000 mm, Y0 = 0.5000 mm 13:59:51 Integrated Voltage Measurement... 14:00:00 Voltage Offset Correction: v_off_ini = 889.670 uV, v_off_fin = 867.278 uV 14:00:00 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -252.831 uVs 14:00:00 Integrated Voltage Measurement... 14:00:09 Voltage Offset Correction: v_off_ini = 882.990 uV, v_off_fin = 900.278 uV 14:00:09 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 248.948 uVs 14:00:09 Integrated Voltage Measurement... 14:00:18 Voltage Offset Correction: v_off_ini = 891.001 uV, v_off_fin = 877.703 uV 14:00:18 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -247.149 uVs 14:00:18 Integrated Voltage Measurement... 14:00:27 Voltage Offset Correction: v_off_ini = 891.326 uV, v_off_fin = 887.968 uV 14:00:27 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 240.328 uVs 14:00:27 Integrated Voltage Measurement... 14:00:36 Voltage Offset Correction: v_off_ini = 894.880 uV, v_off_fin = 880.360 uV 14:00:36 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -257.327 uVs 14:00:36 Integrated Voltage Measurement... 14:00:45 Voltage Offset Correction: v_off_ini = 890.359 uV, v_off_fin = 912.869 uV 14:00:45 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 244.874 uVs 14:00:49 I1Y Average and RMS of Measurements VT = -248.576 +- 3.422 uVs I1Y = -49.713 +- 0.683 uTm 14:00:52 Wire Position: X0 = 0.0000 mm, Y0 = 0.5000 mm 14:00:55 Integrated Voltage Measurement... 14:01:03 Voltage Offset Correction: v_off_ini = 905.610 uV, v_off_fin = 900.892 uV 14:01:04 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -69.407 uVs 14:01:04 Integrated Voltage Measurement... 14:01:13 Voltage Offset Correction: v_off_ini = 900.906 uV, v_off_fin = 912.607 uV 14:01:13 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 77.553 uVs 14:01:13 Integrated Voltage Measurement... 14:01:22 Voltage Offset Correction: v_off_ini = 898.131 uV, v_off_fin = 874.267 uV 14:01:22 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -81.282 uVs 14:01:22 Integrated Voltage Measurement... 14:01:31 Voltage Offset Correction: v_off_ini = 868.401 uV, v_off_fin = 906.093 uV 14:01:31 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 99.880 uVs 14:01:31 Integrated Voltage Measurement... 14:01:40 Voltage Offset Correction: v_off_ini = 911.935 uV, v_off_fin = 888.083 uV 14:01:40 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -65.860 uVs 14:01:40 Integrated Voltage Measurement... 14:01:49 Voltage Offset Correction: v_off_ini = 864.642 uV, v_off_fin = 887.163 uV 14:01:49 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 94.286 uVs 14:01:51 I2X Average and RMS of Measurements VT = -81.378 +- 7.043 uVs I2X = 7.487 +- 0.648 uTm^2 14:01:54 Wire Position: X0 = 0.0000 mm, Y0 = 0.5000 mm 14:01:56 Integrated Voltage Measurement... 14:02:05 Voltage Offset Correction: v_off_ini = 890.800 uV, v_off_fin = 917.055 uV 14:02:05 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -320.003 uVs 14:02:05 Integrated Voltage Measurement... 14:02:14 Voltage Offset Correction: v_off_ini = 904.680 uV, v_off_fin = 904.607 uV 14:02:14 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 312.272 uVs 14:02:14 Integrated Voltage Measurement... 14:02:23 Voltage Offset Correction: v_off_ini = 909.227 uV, v_off_fin = 900.384 uV 14:02:23 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -326.730 uVs 14:02:23 Integrated Voltage Measurement... 14:02:32 Voltage Offset Correction: v_off_ini = 900.930 uV, v_off_fin = 892.200 uV 14:02:32 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 345.872 uVs 14:02:33 Integrated Voltage Measurement... 14:02:41 Voltage Offset Correction: v_off_ini = 904.103 uV, v_off_fin = 899.941 uV 14:02:42 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -322.550 uVs 14:02:42 Integrated Voltage Measurement... 14:02:51 Voltage Offset Correction: v_off_ini = 906.305 uV, v_off_fin = 897.626 uV 14:02:51 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 318.541 uVs 14:02:53 I2Y Average and RMS of Measurements VT = -324.328 +- 8.655 uVs I2Y = -29.838 +- 0.796 uTm^2 14:02:57 Wire Position: X0 = 0.0000 mm, Y0 = 1.0000 mm 14:03:01 Integrated Voltage Measurement... 14:03:10 Voltage Offset Correction: v_off_ini = 899.917 uV, v_off_fin = 889.737 uV 14:03:10 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -64.388 uVs 14:03:10 Integrated Voltage Measurement... 14:03:19 Voltage Offset Correction: v_off_ini = 892.454 uV, v_off_fin = 900.370 uV 14:03:19 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 78.079 uVs 14:03:19 Integrated Voltage Measurement... 14:03:28 Voltage Offset Correction: v_off_ini = 896.524 uV, v_off_fin = 880.205 uV 14:03:28 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -56.045 uVs 14:03:28 Integrated Voltage Measurement... 14:03:37 Voltage Offset Correction: v_off_ini = 852.783 uV, v_off_fin = 865.643 uV 14:03:37 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 56.806 uVs 14:03:37 Integrated Voltage Measurement... 14:03:46 Voltage Offset Correction: v_off_ini = 871.410 uV, v_off_fin = 861.750 uV 14:03:46 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -49.700 uVs 14:03:47 Integrated Voltage Measurement... 14:03:55 Voltage Offset Correction: v_off_ini = 882.333 uV, v_off_fin = 869.808 uV 14:03:56 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 54.403 uVs 14:04:00 I1X Average and RMS of Measurements VT = -59.903 +- 8.208 uVs I1X = 11.981 +- 1.642 uTm 14:04:02 Wire Position: X0 = 0.0000 mm, Y0 = 1.0000 mm 14:04:07 Integrated Voltage Measurement... 14:04:15 Voltage Offset Correction: v_off_ini = 878.497 uV, v_off_fin = 866.117 uV 14:04:15 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -232.623 uVs 14:04:16 Integrated Voltage Measurement... 14:04:24 Voltage Offset Correction: v_off_ini = 874.401 uV, v_off_fin = 895.682 uV 14:04:25 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 257.924 uVs 14:04:25 Integrated Voltage Measurement... 14:04:34 Voltage Offset Correction: v_off_ini = 887.355 uV, v_off_fin = 853.804 uV 14:04:34 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -224.341 uVs 14:04:34 Integrated Voltage Measurement... 14:04:43 Voltage Offset Correction: v_off_ini = 865.069 uV, v_off_fin = 881.179 uV 14:04:43 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 275.641 uVs 14:04:43 Integrated Voltage Measurement... 14:04:52 Voltage Offset Correction: v_off_ini = 868.682 uV, v_off_fin = 838.598 uV 14:04:52 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5002 mm dXY = 0.5001 mm VT = -260.918 uVs 14:04:52 Integrated Voltage Measurement... 14:05:01 Voltage Offset Correction: v_off_ini = 860.412 uV, v_off_fin = 862.075 uV 14:05:01 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 272.580 uVs 14:05:05 I1Y Average and RMS of Measurements VT = -254.005 +- 9.215 uVs I1Y = -50.798 +- 1.840 uTm 14:05:07 Wire Position: X0 = 0.0000 mm, Y0 = 1.0000 mm 14:05:10 Integrated Voltage Measurement... 14:05:19 Voltage Offset Correction: v_off_ini = 878.816 uV, v_off_fin = 877.636 uV 14:05:19 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -72.795 uVs 14:05:19 Integrated Voltage Measurement... 14:05:28 Voltage Offset Correction: v_off_ini = 889.827 uV, v_off_fin = 892.898 uV 14:05:28 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 65.459 uVs 14:05:28 Integrated Voltage Measurement... 14:05:37 Voltage Offset Correction: v_off_ini = 855.554 uV, v_off_fin = 863.738 uV 14:05:37 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -72.236 uVs 14:05:37 Integrated Voltage Measurement... 14:05:46 Voltage Offset Correction: v_off_ini = 872.306 uV, v_off_fin = 872.532 uV 14:05:46 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 71.482 uVs 14:05:46 Integrated Voltage Measurement... 14:05:55 Voltage Offset Correction: v_off_ini = 881.753 uV, v_off_fin = 859.196 uV 14:05:55 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -73.774 uVs 14:05:55 Integrated Voltage Measurement... 14:06:04 Voltage Offset Correction: v_off_ini = 883.113 uV, v_off_fin = 896.911 uV 14:06:04 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 68.025 uVs 14:06:07 I2X Average and RMS of Measurements VT = -70.629 +- 1.132 uVs I2X = 6.498 +- 0.104 uTm^2 14:06:09 Wire Position: X0 = 0.0000 mm, Y0 = 1.0000 mm 14:06:12 Integrated Voltage Measurement... 14:06:21 Voltage Offset Correction: v_off_ini = 867.700 uV, v_off_fin = 868.730 uV 14:06:21 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -315.343 uVs 14:06:21 Integrated Voltage Measurement... 14:06:30 Voltage Offset Correction: v_off_ini = 867.008 uV, v_off_fin = 868.825 uV 14:06:30 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 326.128 uVs 14:06:30 Integrated Voltage Measurement... 14:06:39 Voltage Offset Correction: v_off_ini = 856.626 uV, v_off_fin = 869.738 uV 14:06:39 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -326.323 uVs 14:06:39 Integrated Voltage Measurement... 14:06:48 Voltage Offset Correction: v_off_ini = 856.299 uV, v_off_fin = 856.099 uV 14:06:48 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 332.486 uVs 14:06:48 Integrated Voltage Measurement... 14:06:57 Voltage Offset Correction: v_off_ini = 848.680 uV, v_off_fin = 847.102 uV 14:06:57 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -328.783 uVs 14:06:57 Integrated Voltage Measurement... 14:07:06 Voltage Offset Correction: v_off_ini = 850.234 uV, v_off_fin = 858.663 uV 14:07:06 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 329.019 uVs 14:07:08 I2Y Average and RMS of Measurements VT = -326.347 +- 3.973 uVs I2Y = -30.024 +- 0.366 uTm^2 14:07:24 Wire Position: X0 = -2.0000 mm, Y0 = 0.0000 mm 14:07:29 Integrated Voltage Measurement... 14:07:38 Voltage Offset Correction: v_off_ini = 854.905 uV, v_off_fin = 853.119 uV 14:07:38 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -68.302 uVs 14:07:38 Integrated Voltage Measurement... 14:07:47 Voltage Offset Correction: v_off_ini = 878.303 uV, v_off_fin = 862.004 uV 14:07:47 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 57.458 uVs 14:07:47 Integrated Voltage Measurement... 14:07:56 Voltage Offset Correction: v_off_ini = 871.172 uV, v_off_fin = 878.685 uV 14:07:56 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -80.060 uVs 14:07:56 Integrated Voltage Measurement... 14:08:05 Voltage Offset Correction: v_off_ini = 854.162 uV, v_off_fin = 862.621 uV 14:08:05 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 66.139 uVs 14:08:05 Integrated Voltage Measurement... 14:08:14 Voltage Offset Correction: v_off_ini = 890.024 uV, v_off_fin = 871.730 uV 14:08:14 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -55.242 uVs 14:08:14 Integrated Voltage Measurement... 14:08:23 Voltage Offset Correction: v_off_ini = 846.382 uV, v_off_fin = 827.987 uV 14:08:23 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 108.368 uVs 14:08:27 I1X Average and RMS of Measurements VT = -72.595 +- 7.734 uVs I1X = 14.519 +- 1.547 uTm 14:08:29 Wire Position: X0 = -2.0000 mm, Y0 = 0.0000 mm 14:08:34 Integrated Voltage Measurement... 14:08:43 Voltage Offset Correction: v_off_ini = 841.676 uV, v_off_fin = 839.266 uV 14:08:43 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -246.829 uVs 14:08:43 Integrated Voltage Measurement... 14:08:52 Voltage Offset Correction: v_off_ini = 851.150 uV, v_off_fin = 860.021 uV 14:08:52 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 252.689 uVs 14:08:52 Integrated Voltage Measurement... 14:09:01 Voltage Offset Correction: v_off_ini = 862.892 uV, v_off_fin = 852.248 uV 14:09:01 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -261.227 uVs 14:09:01 Integrated Voltage Measurement... 14:09:10 Voltage Offset Correction: v_off_ini = 850.642 uV, v_off_fin = 859.643 uV 14:09:10 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 269.601 uVs 14:09:10 Integrated Voltage Measurement... 14:09:19 Voltage Offset Correction: v_off_ini = 847.415 uV, v_off_fin = 843.175 uV 14:09:19 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -265.863 uVs 14:09:19 Integrated Voltage Measurement... 14:09:28 Voltage Offset Correction: v_off_ini = 844.409 uV, v_off_fin = 859.650 uV 14:09:28 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 251.150 uVs 14:09:32 I1Y Average and RMS of Measurements VT = -257.893 +- 6.406 uVs I1Y = -51.579 +- 1.281 uTm 14:09:37 Wire Position: X0 = -2.0000 mm, Y0 = 0.0000 mm 14:09:40 Integrated Voltage Measurement... 14:09:49 Voltage Offset Correction: v_off_ini = 832.098 uV, v_off_fin = 837.994 uV 14:09:49 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -95.472 uVs 14:09:49 Integrated Voltage Measurement... 14:09:58 Voltage Offset Correction: v_off_ini = 843.953 uV, v_off_fin = 853.639 uV 14:09:58 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 104.014 uVs 14:09:58 Integrated Voltage Measurement... 14:10:07 Voltage Offset Correction: v_off_ini = 851.524 uV, v_off_fin = 835.065 uV 14:10:07 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -87.850 uVs 14:10:07 Integrated Voltage Measurement... 14:10:16 Voltage Offset Correction: v_off_ini = 840.097 uV, v_off_fin = 844.857 uV 14:10:16 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 84.147 uVs 14:10:16 Integrated Voltage Measurement... 14:10:25 Voltage Offset Correction: v_off_ini = 836.782 uV, v_off_fin = 841.189 uV 14:10:25 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -70.899 uVs 14:10:25 Integrated Voltage Measurement... 14:10:34 Voltage Offset Correction: v_off_ini = 856.256 uV, v_off_fin = 873.131 uV 14:10:34 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 71.466 uVs 14:10:36 I2X Average and RMS of Measurements VT = -85.641 +- 11.663 uVs I2X = 7.879 +- 1.073 uTm^2 14:10:41 Wire Position: X0 = -2.0000 mm, Y0 = 0.0000 mm 14:10:44 Integrated Voltage Measurement... 14:10:53 Voltage Offset Correction: v_off_ini = 812.049 uV, v_off_fin = 836.762 uV 14:10:53 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -323.247 uVs 14:10:53 Integrated Voltage Measurement... 14:11:02 Voltage Offset Correction: v_off_ini = 844.212 uV, v_off_fin = 847.671 uV 14:11:02 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 330.904 uVs 14:11:02 Integrated Voltage Measurement... 14:11:11 Voltage Offset Correction: v_off_ini = 857.053 uV, v_off_fin = 856.139 uV 14:11:11 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -310.813 uVs 14:11:11 Integrated Voltage Measurement... 14:11:20 Voltage Offset Correction: v_off_ini = 846.859 uV, v_off_fin = 840.261 uV 14:11:20 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 324.649 uVs 14:11:20 Integrated Voltage Measurement... 14:11:29 Voltage Offset Correction: v_off_ini = 819.468 uV, v_off_fin = 827.959 uV 14:11:29 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -313.485 uVs 14:11:29 Integrated Voltage Measurement... 14:11:38 Voltage Offset Correction: v_off_ini = 845.407 uV, v_off_fin = 871.245 uV 14:11:38 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 354.759 uVs 14:11:40 I2Y Average and RMS of Measurements VT = -326.310 +- 6.714 uVs I2Y = -30.020 +- 0.618 uTm^2 14:11:45 Wire Position: X0 = -1.0000 mm, Y0 = 0.0000 mm 14:11:49 Integrated Voltage Measurement... 14:11:58 Voltage Offset Correction: v_off_ini = 871.195 uV, v_off_fin = 830.156 uV 14:11:58 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -28.616 uVs 14:11:58 Integrated Voltage Measurement... 14:12:07 Voltage Offset Correction: v_off_ini = 844.890 uV, v_off_fin = 849.381 uV 14:12:07 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 71.451 uVs 14:12:07 Integrated Voltage Measurement... 14:12:16 Voltage Offset Correction: v_off_ini = 865.222 uV, v_off_fin = 859.876 uV 14:12:16 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -65.661 uVs 14:12:16 Integrated Voltage Measurement... 14:12:25 Voltage Offset Correction: v_off_ini = 855.606 uV, v_off_fin = 849.057 uV 14:12:25 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 74.008 uVs 14:12:25 Integrated Voltage Measurement... 14:12:34 Voltage Offset Correction: v_off_ini = 876.070 uV, v_off_fin = 886.008 uV 14:12:34 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -81.147 uVs 14:12:34 Integrated Voltage Measurement... 14:12:43 Voltage Offset Correction: v_off_ini = 884.645 uV, v_off_fin = 891.950 uV 14:12:43 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 53.730 uVs 14:12:48 I1X Average and RMS of Measurements VT = -62.436 +- 8.824 uVs I1X = 12.487 +- 1.765 uTm 14:12:50 Wire Position: X0 = -1.0000 mm, Y0 = 0.0000 mm 14:12:54 Integrated Voltage Measurement... 14:13:03 Voltage Offset Correction: v_off_ini = 867.169 uV, v_off_fin = 846.161 uV 14:13:03 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -250.463 uVs 14:13:03 Integrated Voltage Measurement... 14:13:12 Voltage Offset Correction: v_off_ini = 880.762 uV, v_off_fin = 893.800 uV 14:13:12 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 239.131 uVs 14:13:12 Integrated Voltage Measurement... 14:13:21 Voltage Offset Correction: v_off_ini = 882.315 uV, v_off_fin = 860.809 uV 14:13:21 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -280.366 uVs 14:13:21 Integrated Voltage Measurement... 14:13:30 Voltage Offset Correction: v_off_ini = 877.285 uV, v_off_fin = 873.214 uV 14:13:30 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 244.054 uVs 14:13:30 Integrated Voltage Measurement... 14:13:39 Voltage Offset Correction: v_off_ini = 847.087 uV, v_off_fin = 849.694 uV 14:13:39 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -248.079 uVs 14:13:39 Integrated Voltage Measurement... 14:13:48 Voltage Offset Correction: v_off_ini = 863.915 uV, v_off_fin = 872.428 uV 14:13:48 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 240.791 uVs 14:13:52 I1Y Average and RMS of Measurements VT = -250.481 +- 8.295 uVs I1Y = -50.096 +- 1.659 uTm 14:13:57 Wire Position: X0 = -1.0000 mm, Y0 = 0.0000 mm 14:14:00 Integrated Voltage Measurement... 14:14:09 Voltage Offset Correction: v_off_ini = 866.862 uV, v_off_fin = 873.987 uV 14:14:09 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -72.004 uVs 14:14:09 Integrated Voltage Measurement... 14:14:18 Voltage Offset Correction: v_off_ini = 874.828 uV, v_off_fin = 878.518 uV 14:14:18 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 104.866 uVs 14:14:18 Integrated Voltage Measurement... 14:14:27 Voltage Offset Correction: v_off_ini = 854.114 uV, v_off_fin = 860.390 uV 14:14:27 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -78.933 uVs 14:14:27 Integrated Voltage Measurement... 14:14:36 Voltage Offset Correction: v_off_ini = 879.234 uV, v_off_fin = 868.377 uV 14:14:36 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 61.324 uVs 14:14:36 Integrated Voltage Measurement... 14:14:45 Voltage Offset Correction: v_off_ini = 868.470 uV, v_off_fin = 875.355 uV 14:14:45 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -73.223 uVs 14:14:45 Integrated Voltage Measurement... 14:14:54 Voltage Offset Correction: v_off_ini = 858.657 uV, v_off_fin = 870.340 uV 14:14:54 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 72.074 uVs 14:14:57 I2X Average and RMS of Measurements VT = -77.071 +- 8.101 uVs I2X = 7.091 +- 0.745 uTm^2 14:15:02 Wire Position: X0 = -1.0000 mm, Y0 = 0.0000 mm 14:15:04 Integrated Voltage Measurement... 14:15:13 Voltage Offset Correction: v_off_ini = 891.139 uV, v_off_fin = 894.598 uV 14:15:13 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -351.617 uVs 14:15:13 Integrated Voltage Measurement... 14:15:22 Voltage Offset Correction: v_off_ini = 876.478 uV, v_off_fin = 883.169 uV 14:15:22 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 320.329 uVs 14:15:22 Integrated Voltage Measurement... 14:15:31 Voltage Offset Correction: v_off_ini = 856.453 uV, v_off_fin = 881.316 uV 14:15:31 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -340.714 uVs 14:15:31 Integrated Voltage Measurement... 14:15:40 Voltage Offset Correction: v_off_ini = 895.156 uV, v_off_fin = 890.710 uV 14:15:40 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 321.701 uVs 14:15:40 Integrated Voltage Measurement... 14:15:49 Voltage Offset Correction: v_off_ini = 877.185 uV, v_off_fin = 869.497 uV 14:15:49 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -340.064 uVs 14:15:49 Integrated Voltage Measurement... 14:15:58 Voltage Offset Correction: v_off_ini = 887.190 uV, v_off_fin = 859.358 uV 14:15:58 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 343.719 uVs 14:16:01 I2Y Average and RMS of Measurements VT = -336.357 +- 4.370 uVs I2Y = -30.945 +- 0.402 uTm^2 14:16:05 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 14:16:10 Integrated Voltage Measurement... 14:16:19 Voltage Offset Correction: v_off_ini = 872.106 uV, v_off_fin = 880.377 uV 14:16:19 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -78.892 uVs 14:16:19 Integrated Voltage Measurement... 14:16:28 Voltage Offset Correction: v_off_ini = 876.360 uV, v_off_fin = 879.321 uV 14:16:28 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 58.775 uVs 14:16:28 Integrated Voltage Measurement... 14:16:37 Voltage Offset Correction: v_off_ini = 894.961 uV, v_off_fin = 885.808 uV 14:16:37 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -93.538 uVs 14:16:37 Integrated Voltage Measurement... 14:16:46 Voltage Offset Correction: v_off_ini = 873.960 uV, v_off_fin = 858.753 uV 14:16:46 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 57.687 uVs 14:16:46 Integrated Voltage Measurement... 14:16:55 Voltage Offset Correction: v_off_ini = 876.353 uV, v_off_fin = 876.097 uV 14:16:55 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -63.570 uVs 14:16:55 Integrated Voltage Measurement... 14:17:04 Voltage Offset Correction: v_off_ini = 864.131 uV, v_off_fin = 866.944 uV 14:17:04 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 27.676 uVs 14:17:08 I1X Average and RMS of Measurements VT = -63.356 +- 12.841 uVs I1X = 12.671 +- 2.568 uTm 14:17:11 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 14:17:15 Integrated Voltage Measurement... 14:17:24 Voltage Offset Correction: v_off_ini = 864.354 uV, v_off_fin = 849.361 uV 14:17:24 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -262.766 uVs 14:17:24 Integrated Voltage Measurement... 14:17:33 Voltage Offset Correction: v_off_ini = 887.812 uV, v_off_fin = 882.895 uV 14:17:33 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 229.964 uVs 14:17:33 Integrated Voltage Measurement... 14:17:42 Voltage Offset Correction: v_off_ini = 882.672 uV, v_off_fin = 879.594 uV 14:17:42 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -238.825 uVs 14:17:42 Integrated Voltage Measurement... 14:17:51 Voltage Offset Correction: v_off_ini = 869.717 uV, v_off_fin = 891.853 uV 14:17:51 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 248.745 uVs 14:17:51 Integrated Voltage Measurement... 14:18:00 Voltage Offset Correction: v_off_ini = 873.608 uV, v_off_fin = 871.268 uV 14:18:00 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -233.425 uVs 14:18:00 Integrated Voltage Measurement... 14:18:09 Voltage Offset Correction: v_off_ini = 878.642 uV, v_off_fin = 895.807 uV 14:18:09 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 247.242 uVs 14:18:13 I1Y Average and RMS of Measurements VT = -243.495 +- 2.471 uVs I1Y = -48.699 +- 0.494 uTm 14:18:18 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 14:18:21 Integrated Voltage Measurement... 14:18:30 Voltage Offset Correction: v_off_ini = 901.303 uV, v_off_fin = 877.776 uV 14:18:30 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -80.666 uVs 14:18:30 Integrated Voltage Measurement... 14:18:39 Voltage Offset Correction: v_off_ini = 891.654 uV, v_off_fin = 922.611 uV 14:18:39 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 74.204 uVs 14:18:39 Integrated Voltage Measurement... 14:18:48 Voltage Offset Correction: v_off_ini = 898.102 uV, v_off_fin = 903.715 uV 14:18:48 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -106.327 uVs 14:18:48 Integrated Voltage Measurement... 14:18:57 Voltage Offset Correction: v_off_ini = 905.724 uV, v_off_fin = 886.369 uV 14:18:57 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 85.596 uVs 14:18:57 Integrated Voltage Measurement... 14:19:06 Voltage Offset Correction: v_off_ini = 892.262 uV, v_off_fin = 870.586 uV 14:19:06 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -65.542 uVs 14:19:06 Integrated Voltage Measurement... 14:19:15 Voltage Offset Correction: v_off_ini = 901.309 uV, v_off_fin = 935.443 uV 14:19:15 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 71.351 uVs 14:19:17 I2X Average and RMS of Measurements VT = -80.614 +- 11.456 uVs I2X = 7.417 +- 1.054 uTm^2 14:19:22 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 14:19:25 Integrated Voltage Measurement... 14:19:34 Voltage Offset Correction: v_off_ini = 898.379 uV, v_off_fin = 902.114 uV 14:19:34 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -329.282 uVs 14:19:34 Integrated Voltage Measurement... 14:19:43 Voltage Offset Correction: v_off_ini = 911.787 uV, v_off_fin = 900.953 uV 14:19:43 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 314.030 uVs 14:19:43 Integrated Voltage Measurement... 14:19:52 Voltage Offset Correction: v_off_ini = 903.102 uV, v_off_fin = 896.382 uV 14:19:52 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -334.315 uVs 14:19:52 Integrated Voltage Measurement... 14:20:01 Voltage Offset Correction: v_off_ini = 891.893 uV, v_off_fin = 889.343 uV 14:20:01 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 322.151 uVs 14:20:01 Integrated Voltage Measurement... 14:20:10 Voltage Offset Correction: v_off_ini = 915.060 uV, v_off_fin = 900.584 uV 14:20:10 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -334.990 uVs 14:20:10 Integrated Voltage Measurement... 14:20:19 Voltage Offset Correction: v_off_ini = 888.024 uV, v_off_fin = 898.848 uV 14:20:19 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 331.488 uVs 14:20:22 I2Y Average and RMS of Measurements VT = -327.709 +- 4.743 uVs I2Y = -30.149 +- 0.436 uTm^2 14:20:26 Wire Position: X0 = 1.0000 mm, Y0 = 0.0000 mm 14:20:30 Integrated Voltage Measurement... 14:20:39 Voltage Offset Correction: v_off_ini = 895.178 uV, v_off_fin = 883.760 uV 14:20:39 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -83.004 uVs 14:20:39 Integrated Voltage Measurement... 14:20:48 Voltage Offset Correction: v_off_ini = 891.233 uV, v_off_fin = 886.947 uV 14:20:48 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 86.391 uVs 14:20:48 Integrated Voltage Measurement... 14:20:57 Voltage Offset Correction: v_off_ini = 893.809 uV, v_off_fin = 880.135 uV 14:20:57 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -70.308 uVs 14:20:58 Integrated Voltage Measurement... 14:21:06 Voltage Offset Correction: v_off_ini = 886.585 uV, v_off_fin = 901.200 uV 14:21:06 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 56.495 uVs 14:21:07 Integrated Voltage Measurement... 14:21:15 Voltage Offset Correction: v_off_ini = 891.455 uV, v_off_fin = 884.082 uV 14:21:16 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -93.839 uVs 14:21:16 Integrated Voltage Measurement... 14:21:24 Voltage Offset Correction: v_off_ini = 876.811 uV, v_off_fin = 895.321 uV 14:21:25 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 59.662 uVs 14:21:29 I1X Average and RMS of Measurements VT = -74.950 +- 8.787 uVs I1X = 14.990 +- 1.758 uTm 14:21:31 Wire Position: X0 = 1.0000 mm, Y0 = 0.0000 mm 14:21:35 Integrated Voltage Measurement... 14:21:44 Voltage Offset Correction: v_off_ini = 902.848 uV, v_off_fin = 892.623 uV 14:21:44 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -255.169 uVs 14:21:44 Integrated Voltage Measurement... 14:21:53 Voltage Offset Correction: v_off_ini = 898.400 uV, v_off_fin = 890.991 uV 14:21:53 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 248.222 uVs 14:21:53 Integrated Voltage Measurement... 14:22:02 Voltage Offset Correction: v_off_ini = 889.329 uV, v_off_fin = 901.684 uV 14:22:02 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -245.838 uVs 14:22:02 Integrated Voltage Measurement... 14:22:11 Voltage Offset Correction: v_off_ini = 897.474 uV, v_off_fin = 903.664 uV 14:22:11 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 234.753 uVs 14:22:11 Integrated Voltage Measurement... 14:22:20 Voltage Offset Correction: v_off_ini = 900.507 uV, v_off_fin = 885.960 uV 14:22:20 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -252.846 uVs 14:22:20 Integrated Voltage Measurement... 14:22:29 Voltage Offset Correction: v_off_ini = 900.983 uV, v_off_fin = 884.369 uV 14:22:29 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 260.975 uVs 14:22:34 I1Y Average and RMS of Measurements VT = -249.634 +- 6.938 uVs I1Y = -49.927 +- 1.388 uTm 14:22:39 Wire Position: X0 = 1.0000 mm, Y0 = 0.0000 mm 14:22:41 Integrated Voltage Measurement... 14:22:50 Voltage Offset Correction: v_off_ini = 877.548 uV, v_off_fin = 872.671 uV 14:22:50 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -45.985 uVs 14:22:50 Integrated Voltage Measurement... 14:22:59 Voltage Offset Correction: v_off_ini = 892.045 uV, v_off_fin = 891.210 uV 14:22:59 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 73.108 uVs 14:22:59 Integrated Voltage Measurement... 14:23:08 Voltage Offset Correction: v_off_ini = 899.166 uV, v_off_fin = 881.466 uV 14:23:08 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -99.813 uVs 14:23:08 Integrated Voltage Measurement... 14:23:17 Voltage Offset Correction: v_off_ini = 884.425 uV, v_off_fin = 873.593 uV 14:23:17 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 90.559 uVs 14:23:17 Integrated Voltage Measurement... 14:23:26 Voltage Offset Correction: v_off_ini = 876.984 uV, v_off_fin = 861.398 uV 14:23:26 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -84.759 uVs 14:23:26 Integrated Voltage Measurement... 14:23:35 Voltage Offset Correction: v_off_ini = 870.576 uV, v_off_fin = 877.172 uV 14:23:35 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 75.283 uVs 14:23:38 I2X Average and RMS of Measurements VT = -78.251 +- 14.603 uVs I2X = 7.199 +- 1.344 uTm^2 14:23:43 Wire Position: X0 = 1.0000 mm, Y0 = 0.0000 mm 14:23:46 Integrated Voltage Measurement... 14:23:54 Voltage Offset Correction: v_off_ini = 875.077 uV, v_off_fin = 868.437 uV 14:23:55 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -303.286 uVs 14:23:55 Integrated Voltage Measurement... 14:24:04 Voltage Offset Correction: v_off_ini = 891.932 uV, v_off_fin = 891.369 uV 14:24:04 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 309.949 uVs 14:24:04 Integrated Voltage Measurement... 14:24:13 Voltage Offset Correction: v_off_ini = 896.637 uV, v_off_fin = 891.474 uV 14:24:13 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -336.622 uVs 14:24:13 Integrated Voltage Measurement... 14:24:22 Voltage Offset Correction: v_off_ini = 879.610 uV, v_off_fin = 870.517 uV 14:24:22 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 293.816 uVs 14:24:22 Integrated Voltage Measurement... 14:24:31 Voltage Offset Correction: v_off_ini = 853.672 uV, v_off_fin = 877.130 uV 14:24:31 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -300.195 uVs 14:24:31 Integrated Voltage Measurement... 14:24:40 Voltage Offset Correction: v_off_ini = 879.400 uV, v_off_fin = 891.423 uV 14:24:40 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 328.017 uVs 14:24:42 I2Y Average and RMS of Measurements VT = -311.981 +- 3.820 uVs I2Y = -28.702 +- 0.351 uTm^2 14:24:47 Wire Position: X0 = 2.0000 mm, Y0 = 0.0000 mm 14:24:51 Integrated Voltage Measurement... 14:25:00 Voltage Offset Correction: v_off_ini = 895.136 uV, v_off_fin = 890.147 uV 14:25:00 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.4998 mm dXY = -0.4999 mm VT = -86.250 uVs 14:25:00 Integrated Voltage Measurement... 14:25:09 Voltage Offset Correction: v_off_ini = 892.007 uV, v_off_fin = 902.513 uV 14:25:09 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 63.255 uVs 14:25:09 Integrated Voltage Measurement... 14:25:18 Voltage Offset Correction: v_off_ini = 883.240 uV, v_off_fin = 898.242 uV 14:25:18 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -47.139 uVs 14:25:18 Integrated Voltage Measurement... 14:25:27 Voltage Offset Correction: v_off_ini = 875.807 uV, v_off_fin = 881.254 uV 14:25:27 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 80.371 uVs 14:25:27 Integrated Voltage Measurement... 14:25:36 Voltage Offset Correction: v_off_ini = 887.351 uV, v_off_fin = 913.485 uV 14:25:36 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -67.926 uVs 14:25:36 Integrated Voltage Measurement... 14:25:45 Voltage Offset Correction: v_off_ini = 900.875 uV, v_off_fin = 908.333 uV 14:25:45 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 73.353 uVs 14:25:49 I1X Average and RMS of Measurements VT = -69.716 +- 4.537 uVs I1X = 13.944 +- 0.908 uTm 14:25:52 Wire Position: X0 = 2.0000 mm, Y0 = 0.0000 mm 14:25:56 Integrated Voltage Measurement... 14:26:05 Voltage Offset Correction: v_off_ini = 907.842 uV, v_off_fin = 898.995 uV 14:26:05 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -222.134 uVs 14:26:05 Integrated Voltage Measurement... 14:26:14 Voltage Offset Correction: v_off_ini = 893.464 uV, v_off_fin = 905.441 uV 14:26:14 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 262.108 uVs 14:26:14 Integrated Voltage Measurement... 14:26:23 Voltage Offset Correction: v_off_ini = 914.561 uV, v_off_fin = 902.949 uV 14:26:23 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -250.040 uVs 14:26:23 Integrated Voltage Measurement... 14:26:32 Voltage Offset Correction: v_off_ini = 897.781 uV, v_off_fin = 904.575 uV 14:26:32 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 237.452 uVs 14:26:32 Integrated Voltage Measurement... 14:26:41 Voltage Offset Correction: v_off_ini = 904.402 uV, v_off_fin = 876.282 uV 14:26:41 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.4999 mm VT = -245.454 uVs 14:26:41 Integrated Voltage Measurement... 14:26:50 Voltage Offset Correction: v_off_ini = 897.921 uV, v_off_fin = 887.900 uV 14:26:50 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 251.357 uVs 14:26:54 I1Y Average and RMS of Measurements VT = -244.758 +- 2.663 uVs I1Y = -48.952 +- 0.534 uTm 14:26:59 Wire Position: X0 = 2.0000 mm, Y0 = 0.0000 mm 14:27:02 Integrated Voltage Measurement... 14:27:11 Voltage Offset Correction: v_off_ini = 870.609 uV, v_off_fin = 884.948 uV 14:27:11 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -71.884 uVs 14:27:11 Integrated Voltage Measurement... 14:27:20 Voltage Offset Correction: v_off_ini = 888.440 uV, v_off_fin = 890.159 uV 14:27:20 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 74.171 uVs 14:27:20 Integrated Voltage Measurement... 14:27:29 Voltage Offset Correction: v_off_ini = 882.963 uV, v_off_fin = 857.184 uV 14:27:29 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -92.494 uVs 14:27:29 Integrated Voltage Measurement... 14:27:38 Voltage Offset Correction: v_off_ini = 875.256 uV, v_off_fin = 886.985 uV 14:27:38 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 100.317 uVs 14:27:38 Integrated Voltage Measurement... 14:27:47 Voltage Offset Correction: v_off_ini = 895.555 uV, v_off_fin = 869.652 uV 14:27:47 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -100.477 uVs 14:27:47 Integrated Voltage Measurement... 14:27:56 Voltage Offset Correction: v_off_ini = 877.492 uV, v_off_fin = 892.557 uV 14:27:56 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 82.191 uVs 14:27:59 I2X Average and RMS of Measurements VT = -86.922 +- 10.041 uVs I2X = 7.997 +- 0.924 uTm^2 14:28:04 Wire Position: X0 = 2.0000 mm, Y0 = 0.0000 mm 14:28:06 Integrated Voltage Measurement... 14:28:15 Voltage Offset Correction: v_off_ini = 869.718 uV, v_off_fin = 867.755 uV 14:28:15 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -330.320 uVs 14:28:15 Integrated Voltage Measurement... 14:28:24 Voltage Offset Correction: v_off_ini = 873.323 uV, v_off_fin = 892.559 uV 14:28:24 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 306.122 uVs 14:28:24 Integrated Voltage Measurement... 14:28:33 Voltage Offset Correction: v_off_ini = 913.823 uV, v_off_fin = 901.330 uV 14:28:33 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -328.098 uVs 14:28:33 Integrated Voltage Measurement... 14:28:42 Voltage Offset Correction: v_off_ini = 879.711 uV, v_off_fin = 904.770 uV 14:28:42 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 320.607 uVs 14:28:42 Integrated Voltage Measurement... 14:28:51 Voltage Offset Correction: v_off_ini = 893.789 uV, v_off_fin = 883.674 uV 14:28:51 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -308.477 uVs 14:28:51 Integrated Voltage Measurement... 14:29:00 Voltage Offset Correction: v_off_ini = 897.906 uV, v_off_fin = 893.556 uV 14:29:00 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 316.099 uVs 14:29:03 I2Y Average and RMS of Measurements VT = -318.287 +- 4.925 uVs I2Y = -29.282 +- 0.453 uTm^2 14:29:11 Wire Position: X0 = 0.0000 mm, Y0 = -1.0000 mm 14:29:15 Integrated Voltage Measurement... 14:29:24 Voltage Offset Correction: v_off_ini = 863.718 uV, v_off_fin = 889.828 uV 14:29:24 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -85.313 uVs 14:29:24 Integrated Voltage Measurement... 14:29:33 Voltage Offset Correction: v_off_ini = 862.642 uV, v_off_fin = 876.218 uV 14:29:33 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 75.813 uVs 14:29:33 Integrated Voltage Measurement... 14:29:42 Voltage Offset Correction: v_off_ini = 878.027 uV, v_off_fin = 870.094 uV 14:29:42 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -70.124 uVs 14:29:42 Integrated Voltage Measurement... 14:29:51 Voltage Offset Correction: v_off_ini = 875.814 uV, v_off_fin = 885.985 uV 14:29:51 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 106.007 uVs 14:29:51 Integrated Voltage Measurement... 14:30:00 Voltage Offset Correction: v_off_ini = 865.491 uV, v_off_fin = 865.201 uV 14:30:00 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -68.096 uVs 14:30:00 Integrated Voltage Measurement... 14:30:09 Voltage Offset Correction: v_off_ini = 880.068 uV, v_off_fin = 870.404 uV 14:30:09 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 81.719 uVs 14:30:13 I1X Average and RMS of Measurements VT = -81.179 +- 5.389 uVs I1X = 16.236 +- 1.078 uTm 14:30:16 Wire Position: X0 = 0.0000 mm, Y0 = -1.0000 mm 14:30:20 Integrated Voltage Measurement... 14:30:29 Voltage Offset Correction: v_off_ini = 882.600 uV, v_off_fin = 873.878 uV 14:30:29 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -266.601 uVs 14:30:29 Integrated Voltage Measurement... 14:30:38 Voltage Offset Correction: v_off_ini = 900.708 uV, v_off_fin = 901.722 uV 14:30:38 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 246.819 uVs 14:30:38 Integrated Voltage Measurement... 14:30:47 Voltage Offset Correction: v_off_ini = 893.658 uV, v_off_fin = 868.172 uV 14:30:47 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -262.434 uVs 14:30:47 Integrated Voltage Measurement... 14:30:56 Voltage Offset Correction: v_off_ini = 887.715 uV, v_off_fin = 901.906 uV 14:30:56 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 245.824 uVs 14:30:56 Integrated Voltage Measurement... 14:31:05 Voltage Offset Correction: v_off_ini = 886.145 uV, v_off_fin = 866.480 uV 14:31:05 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -249.685 uVs 14:31:05 Integrated Voltage Measurement... 14:31:14 Voltage Offset Correction: v_off_ini = 874.142 uV, v_off_fin = 866.595 uV 14:31:14 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 239.202 uVs 14:31:18 I1Y Average and RMS of Measurements VT = -251.761 +- 5.280 uVs I1Y = -50.352 +- 1.056 uTm 14:31:20 Wire Position: X0 = 0.0000 mm, Y0 = -1.0000 mm 14:31:23 Integrated Voltage Measurement... 14:31:32 Voltage Offset Correction: v_off_ini = 883.378 uV, v_off_fin = 893.124 uV 14:31:32 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -93.383 uVs 14:31:32 Integrated Voltage Measurement... 14:31:41 Voltage Offset Correction: v_off_ini = 894.598 uV, v_off_fin = 868.344 uV 14:31:41 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 105.275 uVs 14:31:41 Integrated Voltage Measurement... 14:31:50 Voltage Offset Correction: v_off_ini = 890.211 uV, v_off_fin = 865.899 uV 14:31:50 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -75.604 uVs 14:31:50 Integrated Voltage Measurement... 14:31:59 Voltage Offset Correction: v_off_ini = 879.287 uV, v_off_fin = 864.107 uV 14:31:59 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 102.782 uVs 14:31:59 Integrated Voltage Measurement... 14:32:08 Voltage Offset Correction: v_off_ini = 863.703 uV, v_off_fin = 862.463 uV 14:32:08 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -115.567 uVs 14:32:08 Integrated Voltage Measurement... 14:32:17 Voltage Offset Correction: v_off_ini = 864.547 uV, v_off_fin = 870.648 uV 14:32:17 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 96.910 uVs 14:32:20 I2X Average and RMS of Measurements VT = -98.254 +- 7.000 uVs I2X = 9.039 +- 0.644 uTm^2 14:32:22 Wire Position: X0 = 0.0000 mm, Y0 = -1.0000 mm 14:32:25 Integrated Voltage Measurement... 14:32:34 Voltage Offset Correction: v_off_ini = 867.808 uV, v_off_fin = 878.129 uV 14:32:34 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -339.124 uVs 14:32:34 Integrated Voltage Measurement... 14:32:43 Voltage Offset Correction: v_off_ini = 857.980 uV, v_off_fin = 881.470 uV 14:32:43 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 340.219 uVs 14:32:43 Integrated Voltage Measurement... 14:32:52 Voltage Offset Correction: v_off_ini = 872.231 uV, v_off_fin = 863.184 uV 14:32:52 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -345.864 uVs 14:32:52 Integrated Voltage Measurement... 14:33:01 Voltage Offset Correction: v_off_ini = 871.470 uV, v_off_fin = 879.699 uV 14:33:01 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 317.511 uVs 14:33:01 Integrated Voltage Measurement... 14:33:10 Voltage Offset Correction: v_off_ini = 870.911 uV, v_off_fin = 874.221 uV 14:33:10 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -316.914 uVs 14:33:10 Integrated Voltage Measurement... 14:33:19 Voltage Offset Correction: v_off_ini = 878.553 uV, v_off_fin = 870.542 uV 14:33:19 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 319.144 uVs 14:33:21 I2Y Average and RMS of Measurements VT = -329.796 +- 8.936 uVs I2Y = -30.341 +- 0.822 uTm^2 14:33:25 Wire Position: X0 = 0.0000 mm, Y0 = -0.5000 mm 14:33:30 Integrated Voltage Measurement... 14:33:39 Voltage Offset Correction: v_off_ini = 856.000 uV, v_off_fin = 865.832 uV 14:33:39 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -61.751 uVs 14:33:39 Integrated Voltage Measurement... 14:33:48 Voltage Offset Correction: v_off_ini = 852.562 uV, v_off_fin = 843.323 uV 14:33:48 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 80.756 uVs 14:33:48 Integrated Voltage Measurement... 14:33:57 Voltage Offset Correction: v_off_ini = 851.978 uV, v_off_fin = 849.975 uV 14:33:57 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -59.583 uVs 14:33:57 Integrated Voltage Measurement... 14:34:06 Voltage Offset Correction: v_off_ini = 852.973 uV, v_off_fin = 849.023 uV 14:34:06 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 84.078 uVs 14:34:06 Integrated Voltage Measurement... 14:34:15 Voltage Offset Correction: v_off_ini = 838.715 uV, v_off_fin = 868.562 uV 14:34:15 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -41.130 uVs 14:34:15 Integrated Voltage Measurement... 14:34:24 Voltage Offset Correction: v_off_ini = 841.984 uV, v_off_fin = 862.196 uV 14:34:24 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 64.589 uVs 14:34:28 I1X Average and RMS of Measurements VT = -65.314 +- 8.810 uVs I1X = 13.063 +- 1.762 uTm 14:34:30 Wire Position: X0 = 0.0000 mm, Y0 = -0.5000 mm 14:34:35 Integrated Voltage Measurement... 14:34:44 Voltage Offset Correction: v_off_ini = 853.037 uV, v_off_fin = 856.503 uV 14:34:44 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -258.958 uVs 14:34:44 Integrated Voltage Measurement... 14:34:53 Voltage Offset Correction: v_off_ini = 847.906 uV, v_off_fin = 871.339 uV 14:34:53 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 255.913 uVs 14:34:53 Integrated Voltage Measurement... 14:35:02 Voltage Offset Correction: v_off_ini = 850.418 uV, v_off_fin = 850.390 uV 14:35:02 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -240.270 uVs 14:35:02 Integrated Voltage Measurement... 14:35:11 Voltage Offset Correction: v_off_ini = 850.588 uV, v_off_fin = 885.633 uV 14:35:11 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 257.101 uVs 14:35:11 Integrated Voltage Measurement... 14:35:20 Voltage Offset Correction: v_off_ini = 870.284 uV, v_off_fin = 868.356 uV 14:35:20 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -255.486 uVs 14:35:20 Integrated Voltage Measurement... 14:35:29 Voltage Offset Correction: v_off_ini = 852.570 uV, v_off_fin = 872.175 uV 14:35:29 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 236.405 uVs 14:35:33 I1Y Average and RMS of Measurements VT = -250.689 +- 4.900 uVs I1Y = -50.138 +- 0.980 uTm 14:35:35 Wire Position: X0 = 0.0000 mm, Y0 = -0.5000 mm 14:35:38 Integrated Voltage Measurement... 14:35:47 Voltage Offset Correction: v_off_ini = 868.920 uV, v_off_fin = 860.598 uV 14:35:47 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -97.433 uVs 14:35:47 Integrated Voltage Measurement... 14:35:56 Voltage Offset Correction: v_off_ini = 884.941 uV, v_off_fin = 896.000 uV 14:35:56 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 87.475 uVs 14:35:56 Integrated Voltage Measurement... 14:36:05 Voltage Offset Correction: v_off_ini = 859.596 uV, v_off_fin = 867.583 uV 14:36:05 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -68.694 uVs 14:36:05 Integrated Voltage Measurement... 14:36:14 Voltage Offset Correction: v_off_ini = 880.697 uV, v_off_fin = 868.184 uV 14:36:14 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 68.482 uVs 14:36:14 Integrated Voltage Measurement... 14:36:23 Voltage Offset Correction: v_off_ini = 875.898 uV, v_off_fin = 856.353 uV 14:36:23 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -86.528 uVs 14:36:23 Integrated Voltage Measurement... 14:36:32 Voltage Offset Correction: v_off_ini = 861.888 uV, v_off_fin = 859.108 uV 14:36:32 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 96.260 uVs 14:36:34 I2X Average and RMS of Measurements VT = -84.145 +- 11.009 uVs I2X = 7.741 +- 1.013 uTm^2 14:36:37 Wire Position: X0 = 0.0000 mm, Y0 = -0.5000 mm 14:36:40 Integrated Voltage Measurement... 14:36:49 Voltage Offset Correction: v_off_ini = 853.811 uV, v_off_fin = 829.389 uV 14:36:49 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -306.983 uVs 14:36:49 Integrated Voltage Measurement... 14:36:58 Voltage Offset Correction: v_off_ini = 837.338 uV, v_off_fin = 847.716 uV 14:36:58 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 344.019 uVs 14:36:58 Integrated Voltage Measurement... 14:37:07 Voltage Offset Correction: v_off_ini = 879.343 uV, v_off_fin = 869.250 uV 14:37:07 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -321.156 uVs 14:37:07 Integrated Voltage Measurement... 14:37:16 Voltage Offset Correction: v_off_ini = 868.990 uV, v_off_fin = 854.859 uV 14:37:16 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 325.281 uVs 14:37:16 Integrated Voltage Measurement... 14:37:25 Voltage Offset Correction: v_off_ini = 876.936 uV, v_off_fin = 860.972 uV 14:37:25 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -315.456 uVs 14:37:25 Integrated Voltage Measurement... 14:37:34 Voltage Offset Correction: v_off_ini = 868.036 uV, v_off_fin = 857.210 uV 14:37:34 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 329.083 uVs 14:37:36 I2Y Average and RMS of Measurements VT = -323.663 +- 1.356 uVs I2Y = -29.777 +- 0.125 uTm^2 14:37:40 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 14:37:45 Integrated Voltage Measurement... 14:37:53 Voltage Offset Correction: v_off_ini = 879.216 uV, v_off_fin = 869.932 uV 14:37:54 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -87.603 uVs 14:37:54 Integrated Voltage Measurement... 14:38:03 Voltage Offset Correction: v_off_ini = 866.650 uV, v_off_fin = 874.185 uV 14:38:03 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 89.509 uVs 14:38:03 Integrated Voltage Measurement... 14:38:12 Voltage Offset Correction: v_off_ini = 852.559 uV, v_off_fin = 844.687 uV 14:38:12 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -57.649 uVs 14:38:12 Integrated Voltage Measurement... 14:38:21 Voltage Offset Correction: v_off_ini = 868.572 uV, v_off_fin = 862.944 uV 14:38:21 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 79.236 uVs 14:38:21 Integrated Voltage Measurement... 14:38:30 Voltage Offset Correction: v_off_ini = 860.129 uV, v_off_fin = 875.135 uV 14:38:30 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -63.476 uVs 14:38:30 Integrated Voltage Measurement... 14:38:39 Voltage Offset Correction: v_off_ini = 877.170 uV, v_off_fin = 870.300 uV 14:38:39 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 69.172 uVs 14:38:43 I1X Average and RMS of Measurements VT = -74.441 +- 10.018 uVs I1X = 14.888 +- 2.004 uTm 14:38:45 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 14:38:50 Integrated Voltage Measurement... 14:38:59 Voltage Offset Correction: v_off_ini = 866.808 uV, v_off_fin = 858.523 uV 14:38:59 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -261.744 uVs 14:38:59 Integrated Voltage Measurement... 14:39:08 Voltage Offset Correction: v_off_ini = 859.963 uV, v_off_fin = 873.216 uV 14:39:08 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 244.505 uVs 14:39:08 Integrated Voltage Measurement... 14:39:17 Voltage Offset Correction: v_off_ini = 847.497 uV, v_off_fin = 848.766 uV 14:39:17 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -246.397 uVs 14:39:17 Integrated Voltage Measurement... 14:39:26 Voltage Offset Correction: v_off_ini = 876.717 uV, v_off_fin = 882.641 uV 14:39:26 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 247.869 uVs 14:39:26 Integrated Voltage Measurement... 14:39:35 Voltage Offset Correction: v_off_ini = 884.638 uV, v_off_fin = 891.035 uV 14:39:35 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -245.425 uVs 14:39:35 Integrated Voltage Measurement... 14:39:44 Voltage Offset Correction: v_off_ini = 903.079 uV, v_off_fin = 922.801 uV 14:39:44 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 241.179 uVs 14:39:48 I1Y Average and RMS of Measurements VT = -247.853 +- 4.043 uVs I1Y = -49.571 +- 0.810 uTm 14:39:50 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 14:39:53 Integrated Voltage Measurement... 14:40:02 Voltage Offset Correction: v_off_ini = 856.544 uV, v_off_fin = 854.057 uV 14:40:02 Integrated Voltage Measurement dXYus = -1.0001 mm, dXYds = 0.0000 mm dXY = -1.0001 mm VT = -66.203 uVs 14:40:02 Integrated Voltage Measurement... 14:40:11 Voltage Offset Correction: v_off_ini = 855.642 uV, v_off_fin = 876.267 uV 14:40:11 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 94.046 uVs 14:40:11 Integrated Voltage Measurement... 14:40:20 Voltage Offset Correction: v_off_ini = 873.877 uV, v_off_fin = 862.835 uV 14:40:20 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -101.532 uVs 14:40:20 Integrated Voltage Measurement... 14:40:29 Voltage Offset Correction: v_off_ini = 883.872 uV, v_off_fin = 863.976 uV 14:40:29 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 74.363 uVs 14:40:29 Integrated Voltage Measurement... 14:40:38 Voltage Offset Correction: v_off_ini = 864.773 uV, v_off_fin = 854.641 uV 14:40:38 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -74.252 uVs 14:40:38 Integrated Voltage Measurement... 14:40:47 Voltage Offset Correction: v_off_ini = 876.747 uV, v_off_fin = 885.076 uV 14:40:47 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 67.602 uVs 14:40:50 I2X Average and RMS of Measurements VT = -79.667 +- 6.956 uVs I2X = 7.329 +- 0.640 uTm^2 14:40:52 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 14:40:55 Integrated Voltage Measurement... 14:41:04 Voltage Offset Correction: v_off_ini = 879.499 uV, v_off_fin = 878.441 uV 14:41:04 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -335.905 uVs 14:41:04 Integrated Voltage Measurement... 14:41:13 Voltage Offset Correction: v_off_ini = 863.273 uV, v_off_fin = 856.854 uV 14:41:13 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 326.996 uVs 14:41:13 Integrated Voltage Measurement... 14:41:22 Voltage Offset Correction: v_off_ini = 871.757 uV, v_off_fin = 870.554 uV 14:41:22 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -331.789 uVs 14:41:22 Integrated Voltage Measurement... 14:41:31 Voltage Offset Correction: v_off_ini = 879.649 uV, v_off_fin = 878.459 uV 14:41:31 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 310.860 uVs 14:41:31 Integrated Voltage Measurement... 14:41:40 Voltage Offset Correction: v_off_ini = 881.851 uV, v_off_fin = 902.839 uV 14:41:40 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -334.460 uVs 14:41:40 Integrated Voltage Measurement... 14:41:49 Voltage Offset Correction: v_off_ini = 892.024 uV, v_off_fin = 877.072 uV 14:41:49 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 308.706 uVs 14:41:51 I2Y Average and RMS of Measurements VT = -324.786 +- 4.714 uVs I2Y = -29.880 +- 0.434 uTm^2 14:41:55 Wire Position: X0 = 0.0000 mm, Y0 = 0.5000 mm 14:42:00 Integrated Voltage Measurement... 14:42:09 Voltage Offset Correction: v_off_ini = 873.360 uV, v_off_fin = 860.848 uV 14:42:09 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -55.482 uVs 14:42:09 Integrated Voltage Measurement... 14:42:18 Voltage Offset Correction: v_off_ini = 845.996 uV, v_off_fin = 863.410 uV 14:42:18 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 51.536 uVs 14:42:18 Integrated Voltage Measurement... 14:42:27 Voltage Offset Correction: v_off_ini = 884.294 uV, v_off_fin = 859.480 uV 14:42:27 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -45.098 uVs 14:42:27 Integrated Voltage Measurement... 14:42:36 Voltage Offset Correction: v_off_ini = 882.008 uV, v_off_fin = 860.237 uV 14:42:36 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 64.925 uVs 14:42:36 Integrated Voltage Measurement... 14:42:45 Voltage Offset Correction: v_off_ini = 866.855 uV, v_off_fin = 876.569 uV 14:42:45 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -51.163 uVs 14:42:45 Integrated Voltage Measurement... 14:42:54 Voltage Offset Correction: v_off_ini = 882.921 uV, v_off_fin = 861.559 uV 14:42:54 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 38.887 uVs 14:42:58 I1X Average and RMS of Measurements VT = -51.182 +- 4.397 uVs I1X = 10.237 +- 0.879 uTm 14:43:01 Wire Position: X0 = 0.0000 mm, Y0 = 0.5000 mm 14:43:05 Integrated Voltage Measurement... 14:43:14 Voltage Offset Correction: v_off_ini = 864.591 uV, v_off_fin = 863.419 uV 14:43:14 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -240.476 uVs 14:43:14 Integrated Voltage Measurement... 14:43:23 Voltage Offset Correction: v_off_ini = 883.665 uV, v_off_fin = 883.041 uV 14:43:23 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 222.438 uVs 14:43:23 Integrated Voltage Measurement... 14:43:32 Voltage Offset Correction: v_off_ini = 906.378 uV, v_off_fin = 877.223 uV 14:43:32 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5002 mm dXY = 0.5001 mm VT = -257.723 uVs 14:43:32 Integrated Voltage Measurement... 14:43:41 Voltage Offset Correction: v_off_ini = 899.227 uV, v_off_fin = 895.043 uV 14:43:41 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 258.904 uVs 14:43:41 Integrated Voltage Measurement... 14:43:50 Voltage Offset Correction: v_off_ini = 875.659 uV, v_off_fin = 871.300 uV 14:43:50 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -246.987 uVs 14:43:50 Integrated Voltage Measurement... 14:43:59 Voltage Offset Correction: v_off_ini = 873.042 uV, v_off_fin = 886.342 uV 14:43:59 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5002 mm dXY = -0.5001 mm VT = 258.274 uVs 14:44:03 I1Y Average and RMS of Measurements VT = -247.467 +- 11.556 uVs I1Y = -49.488 +- 2.308 uTm 14:44:06 Wire Position: X0 = 0.0000 mm, Y0 = 0.5000 mm 14:44:08 Integrated Voltage Measurement... 14:44:17 Voltage Offset Correction: v_off_ini = 872.343 uV, v_off_fin = 850.851 uV 14:44:17 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -108.326 uVs 14:44:17 Integrated Voltage Measurement... 14:44:26 Voltage Offset Correction: v_off_ini = 862.837 uV, v_off_fin = 869.329 uV 14:44:26 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 66.155 uVs 14:44:26 Integrated Voltage Measurement... 14:44:35 Voltage Offset Correction: v_off_ini = 879.346 uV, v_off_fin = 863.495 uV 14:44:35 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -77.922 uVs 14:44:35 Integrated Voltage Measurement... 14:44:44 Voltage Offset Correction: v_off_ini = 861.866 uV, v_off_fin = 876.376 uV 14:44:44 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 82.424 uVs 14:44:44 Integrated Voltage Measurement... 14:44:53 Voltage Offset Correction: v_off_ini = 886.410 uV, v_off_fin = 854.884 uV 14:44:53 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -39.346 uVs 14:44:53 Integrated Voltage Measurement... 14:45:02 Voltage Offset Correction: v_off_ini = 844.766 uV, v_off_fin = 858.700 uV 14:45:02 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 93.428 uVs 14:45:05 I2X Average and RMS of Measurements VT = -77.933 +- 8.659 uVs I2X = 7.170 +- 0.796 uTm^2 14:45:07 Wire Position: X0 = 0.0000 mm, Y0 = 0.5000 mm 14:45:10 Integrated Voltage Measurement... 14:45:19 Voltage Offset Correction: v_off_ini = 875.370 uV, v_off_fin = 871.429 uV 14:45:19 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -340.025 uVs 14:45:19 Integrated Voltage Measurement... 14:45:28 Voltage Offset Correction: v_off_ini = 867.275 uV, v_off_fin = 857.396 uV 14:45:28 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 328.655 uVs 14:45:28 Integrated Voltage Measurement... 14:45:37 Voltage Offset Correction: v_off_ini = 840.307 uV, v_off_fin = 853.920 uV 14:45:37 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -322.740 uVs 14:45:37 Integrated Voltage Measurement... 14:45:46 Voltage Offset Correction: v_off_ini = 869.995 uV, v_off_fin = 855.289 uV 14:45:46 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 325.879 uVs 14:45:46 Integrated Voltage Measurement... 14:45:55 Voltage Offset Correction: v_off_ini = 866.610 uV, v_off_fin = 890.824 uV 14:45:55 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -304.973 uVs 14:45:55 Integrated Voltage Measurement... 14:46:04 Voltage Offset Correction: v_off_ini = 870.938 uV, v_off_fin = 855.131 uV 14:46:04 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 308.082 uVs 14:46:07 I2Y Average and RMS of Measurements VT = -321.726 +- 11.501 uVs I2Y = -29.599 +- 1.058 uTm^2 14:46:11 Wire Position: X0 = 0.0000 mm, Y0 = 1.0000 mm 14:46:15 Integrated Voltage Measurement... 14:46:24 Voltage Offset Correction: v_off_ini = 888.538 uV, v_off_fin = 887.345 uV 14:46:24 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -72.075 uVs 14:46:24 Integrated Voltage Measurement... 14:46:33 Voltage Offset Correction: v_off_ini = 882.030 uV, v_off_fin = 868.119 uV 14:46:33 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 65.289 uVs 14:46:33 Integrated Voltage Measurement... 14:46:42 Voltage Offset Correction: v_off_ini = 867.579 uV, v_off_fin = 891.053 uV 14:46:42 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -68.187 uVs 14:46:42 Integrated Voltage Measurement... 14:46:51 Voltage Offset Correction: v_off_ini = 893.825 uV, v_off_fin = 898.924 uV 14:46:51 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 44.894 uVs 14:46:51 Integrated Voltage Measurement... 14:47:00 Voltage Offset Correction: v_off_ini = 898.823 uV, v_off_fin = 894.273 uV 14:47:00 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -8.708 uVs 14:47:00 Integrated Voltage Measurement... 14:47:09 Voltage Offset Correction: v_off_ini = 874.041 uV, v_off_fin = 877.517 uV 14:47:09 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 72.699 uVs 14:47:13 I1X Average and RMS of Measurements VT = -55.309 +- 11.455 uVs I1X = 11.062 +- 2.291 uTm 14:47:16 Wire Position: X0 = 0.0000 mm, Y0 = 1.0000 mm 14:47:20 Integrated Voltage Measurement... 14:47:29 Voltage Offset Correction: v_off_ini = 893.707 uV, v_off_fin = 854.435 uV 14:47:29 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -221.289 uVs 14:47:29 Integrated Voltage Measurement... 14:47:38 Voltage Offset Correction: v_off_ini = 893.723 uV, v_off_fin = 912.529 uV 14:47:38 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 245.164 uVs 14:47:38 Integrated Voltage Measurement... 14:47:47 Voltage Offset Correction: v_off_ini = 888.795 uV, v_off_fin = 869.955 uV 14:47:47 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5002 mm dXY = 0.5001 mm VT = -249.371 uVs 14:47:47 Integrated Voltage Measurement... 14:47:56 Voltage Offset Correction: v_off_ini = 883.389 uV, v_off_fin = 900.491 uV 14:47:56 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5002 mm dXY = -0.5001 mm VT = 250.977 uVs 14:47:56 Integrated Voltage Measurement... 14:48:05 Voltage Offset Correction: v_off_ini = 883.974 uV, v_off_fin = 886.286 uV 14:48:05 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -267.719 uVs 14:48:05 Integrated Voltage Measurement... 14:48:14 Voltage Offset Correction: v_off_ini = 887.897 uV, v_off_fin = 895.040 uV 14:48:14 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 243.792 uVs 14:48:18 I1Y Average and RMS of Measurements VT = -246.385 +- 9.579 uVs I1Y = -49.274 +- 1.915 uTm 14:48:21 Wire Position: X0 = 0.0000 mm, Y0 = 1.0000 mm 14:48:23 Integrated Voltage Measurement... 14:48:32 Voltage Offset Correction: v_off_ini = 881.261 uV, v_off_fin = 867.556 uV 14:48:32 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -87.052 uVs 14:48:32 Integrated Voltage Measurement... 14:48:41 Voltage Offset Correction: v_off_ini = 879.088 uV, v_off_fin = 870.832 uV 14:48:41 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 76.180 uVs 14:48:41 Integrated Voltage Measurement... 14:48:50 Voltage Offset Correction: v_off_ini = 860.996 uV, v_off_fin = 885.392 uV 14:48:50 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -52.318 uVs 14:48:51 Integrated Voltage Measurement... 14:49:00 Voltage Offset Correction: v_off_ini = 889.075 uV, v_off_fin = 881.025 uV 14:49:00 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 63.053 uVs 14:49:00 Integrated Voltage Measurement... 14:49:09 Voltage Offset Correction: v_off_ini = 882.137 uV, v_off_fin = 886.482 uV 14:49:09 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -67.442 uVs 14:49:09 Integrated Voltage Measurement... 14:49:18 Voltage Offset Correction: v_off_ini = 865.323 uV, v_off_fin = 901.754 uV 14:49:18 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 76.592 uVs 14:49:20 I2X Average and RMS of Measurements VT = -70.440 +- 9.833 uVs I2X = 6.480 +- 0.905 uTm^2 14:49:22 Wire Position: X0 = 0.0000 mm, Y0 = 1.0000 mm 14:49:25 Integrated Voltage Measurement... 14:49:34 Voltage Offset Correction: v_off_ini = 872.715 uV, v_off_fin = 879.273 uV 14:49:34 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -325.559 uVs 14:49:34 Integrated Voltage Measurement... 14:49:43 Voltage Offset Correction: v_off_ini = 889.655 uV, v_off_fin = 880.752 uV 14:49:43 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 327.207 uVs 14:49:43 Integrated Voltage Measurement... 14:49:52 Voltage Offset Correction: v_off_ini = 886.089 uV, v_off_fin = 880.879 uV 14:49:52 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -327.954 uVs 14:49:52 Integrated Voltage Measurement... 14:50:01 Voltage Offset Correction: v_off_ini = 875.700 uV, v_off_fin = 889.094 uV 14:50:01 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 334.474 uVs 14:50:01 Integrated Voltage Measurement... 14:50:10 Voltage Offset Correction: v_off_ini = 880.720 uV, v_off_fin = 875.653 uV 14:50:10 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -342.007 uVs 14:50:10 Integrated Voltage Measurement... 14:50:19 Voltage Offset Correction: v_off_ini = 860.454 uV, v_off_fin = 863.356 uV 14:50:19 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 313.040 uVs 14:50:22 I2Y Average and RMS of Measurements VT = -328.374 +- 2.062 uVs I2Y = -30.210 +- 0.190 uTm^2 14:50:38 Wire Position: X0 = -2.0000 mm, Y0 = 0.0000 mm 14:50:42 Integrated Voltage Measurement... 14:50:51 Voltage Offset Correction: v_off_ini = 851.290 uV, v_off_fin = 863.070 uV 14:50:51 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.4998 mm dXY = -0.4999 mm VT = -79.689 uVs 14:50:51 Integrated Voltage Measurement... 14:51:00 Voltage Offset Correction: v_off_ini = 842.404 uV, v_off_fin = 863.885 uV 14:51:00 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 66.613 uVs 14:51:00 Integrated Voltage Measurement... 14:51:09 Voltage Offset Correction: v_off_ini = 870.608 uV, v_off_fin = 870.683 uV 14:51:09 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -57.463 uVs 14:51:09 Integrated Voltage Measurement... 14:51:18 Voltage Offset Correction: v_off_ini = 865.119 uV, v_off_fin = 873.524 uV 14:51:18 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 46.848 uVs 14:51:18 Integrated Voltage Measurement... 14:51:27 Voltage Offset Correction: v_off_ini = 864.132 uV, v_off_fin = 862.796 uV 14:51:27 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -35.524 uVs 14:51:27 Integrated Voltage Measurement... 14:51:36 Voltage Offset Correction: v_off_ini = 893.198 uV, v_off_fin = 891.594 uV 14:51:36 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 51.822 uVs 14:51:40 I1X Average and RMS of Measurements VT = -56.327 +- 12.391 uVs I1X = 11.266 +- 2.479 uTm 14:51:43 Wire Position: X0 = -2.0000 mm, Y0 = 0.0000 mm 14:51:47 Integrated Voltage Measurement... 14:51:56 Voltage Offset Correction: v_off_ini = 882.724 uV, v_off_fin = 871.755 uV 14:51:56 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -265.928 uVs 14:51:56 Integrated Voltage Measurement... 14:52:05 Voltage Offset Correction: v_off_ini = 887.897 uV, v_off_fin = 881.894 uV 14:52:05 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 279.078 uVs 14:52:05 Integrated Voltage Measurement... 14:52:14 Voltage Offset Correction: v_off_ini = 894.727 uV, v_off_fin = 883.902 uV 14:52:14 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -243.294 uVs 14:52:14 Integrated Voltage Measurement... 14:52:23 Voltage Offset Correction: v_off_ini = 886.489 uV, v_off_fin = 879.680 uV 14:52:23 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 245.690 uVs 14:52:23 Integrated Voltage Measurement... 14:52:32 Voltage Offset Correction: v_off_ini = 877.984 uV, v_off_fin = 868.858 uV 14:52:32 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -255.571 uVs 14:52:32 Integrated Voltage Measurement... 14:52:41 Voltage Offset Correction: v_off_ini = 895.171 uV, v_off_fin = 901.499 uV 14:52:41 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 261.744 uVs 14:52:45 I1Y Average and RMS of Measurements VT = -258.551 +- 11.436 uVs I1Y = -51.710 +- 2.287 uTm 14:52:50 Wire Position: X0 = -2.0000 mm, Y0 = 0.0000 mm 14:52:53 Integrated Voltage Measurement... 14:53:02 Voltage Offset Correction: v_off_ini = 879.766 uV, v_off_fin = 861.313 uV 14:53:02 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -81.749 uVs 14:53:02 Integrated Voltage Measurement... 14:53:11 Voltage Offset Correction: v_off_ini = 878.220 uV, v_off_fin = 894.652 uV 14:53:11 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 86.790 uVs 14:53:11 Integrated Voltage Measurement... 14:53:20 Voltage Offset Correction: v_off_ini = 877.985 uV, v_off_fin = 886.597 uV 14:53:20 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -90.996 uVs 14:53:20 Integrated Voltage Measurement... 14:53:29 Voltage Offset Correction: v_off_ini = 878.179 uV, v_off_fin = 892.919 uV 14:53:29 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 90.426 uVs 14:53:29 Integrated Voltage Measurement... 14:53:38 Voltage Offset Correction: v_off_ini = 876.539 uV, v_off_fin = 878.109 uV 14:53:38 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -65.575 uVs 14:53:38 Integrated Voltage Measurement... 14:53:47 Voltage Offset Correction: v_off_ini = 858.432 uV, v_off_fin = 876.754 uV 14:53:47 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 81.235 uVs 14:53:50 I2X Average and RMS of Measurements VT = -82.795 +- 7.141 uVs I2X = 7.617 +- 0.657 uTm^2 14:53:55 Wire Position: X0 = -2.0000 mm, Y0 = 0.0000 mm 14:53:57 Integrated Voltage Measurement... 14:54:06 Voltage Offset Correction: v_off_ini = 901.808 uV, v_off_fin = 886.664 uV 14:54:06 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -297.118 uVs 14:54:06 Integrated Voltage Measurement... 14:54:15 Voltage Offset Correction: v_off_ini = 878.473 uV, v_off_fin = 891.661 uV 14:54:15 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 348.846 uVs 14:54:16 Integrated Voltage Measurement... 14:54:24 Voltage Offset Correction: v_off_ini = 904.328 uV, v_off_fin = 913.303 uV 14:54:25 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -313.672 uVs 14:54:25 Integrated Voltage Measurement... 14:54:34 Voltage Offset Correction: v_off_ini = 916.197 uV, v_off_fin = 918.041 uV 14:54:34 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 321.791 uVs 14:54:34 Integrated Voltage Measurement... 14:54:43 Voltage Offset Correction: v_off_ini = 911.527 uV, v_off_fin = 897.885 uV 14:54:43 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -332.041 uVs 14:54:43 Integrated Voltage Measurement... 14:54:52 Voltage Offset Correction: v_off_ini = 896.278 uV, v_off_fin = 897.717 uV 14:54:52 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 336.734 uVs 14:54:54 I2Y Average and RMS of Measurements VT = -325.034 +- 6.953 uVs I2Y = -29.903 +- 0.640 uTm^2 14:54:59 Wire Position: X0 = -1.0000 mm, Y0 = 0.0000 mm 14:55:03 Integrated Voltage Measurement... 14:55:12 Voltage Offset Correction: v_off_ini = 892.311 uV, v_off_fin = 911.530 uV 14:55:12 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -67.731 uVs 14:55:12 Integrated Voltage Measurement... 14:55:21 Voltage Offset Correction: v_off_ini = 908.834 uV, v_off_fin = 911.086 uV 14:55:21 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 61.696 uVs 14:55:21 Integrated Voltage Measurement... 14:55:30 Voltage Offset Correction: v_off_ini = 892.141 uV, v_off_fin = 887.181 uV 14:55:30 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -60.617 uVs 14:55:30 Integrated Voltage Measurement... 14:55:39 Voltage Offset Correction: v_off_ini = 891.191 uV, v_off_fin = 886.157 uV 14:55:39 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 42.803 uVs 14:55:39 Integrated Voltage Measurement... 14:55:48 Voltage Offset Correction: v_off_ini = 882.424 uV, v_off_fin = 883.811 uV 14:55:48 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -33.425 uVs 14:55:48 Integrated Voltage Measurement... 14:55:57 Voltage Offset Correction: v_off_ini = 882.721 uV, v_off_fin = 890.441 uV 14:55:57 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 56.032 uVs 14:56:01 I1X Average and RMS of Measurements VT = -53.717 +- 8.282 uVs I1X = 10.743 +- 1.656 uTm 14:56:04 Wire Position: X0 = -1.0000 mm, Y0 = 0.0000 mm 14:56:08 Integrated Voltage Measurement... 14:56:17 Voltage Offset Correction: v_off_ini = 869.357 uV, v_off_fin = 851.540 uV 14:56:17 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -248.868 uVs 14:56:17 Integrated Voltage Measurement... 14:56:26 Voltage Offset Correction: v_off_ini = 868.115 uV, v_off_fin = 883.668 uV 14:56:26 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 255.751 uVs 14:56:26 Integrated Voltage Measurement... 14:56:35 Voltage Offset Correction: v_off_ini = 902.986 uV, v_off_fin = 880.521 uV 14:56:35 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -258.745 uVs 14:56:35 Integrated Voltage Measurement... 14:56:44 Voltage Offset Correction: v_off_ini = 899.340 uV, v_off_fin = 907.762 uV 14:56:44 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 239.603 uVs 14:56:44 Integrated Voltage Measurement... 14:56:53 Voltage Offset Correction: v_off_ini = 886.247 uV, v_off_fin = 866.689 uV 14:56:53 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -263.212 uVs 14:56:53 Integrated Voltage Measurement... 14:57:02 Voltage Offset Correction: v_off_ini = 883.657 uV, v_off_fin = 904.879 uV 14:57:02 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 267.610 uVs 14:57:06 I1Y Average and RMS of Measurements VT = -255.631 +- 7.033 uVs I1Y = -51.127 +- 1.408 uTm 14:57:11 Wire Position: X0 = -1.0000 mm, Y0 = 0.0000 mm 14:57:14 Integrated Voltage Measurement... 14:57:23 Voltage Offset Correction: v_off_ini = 877.315 uV, v_off_fin = 860.391 uV 14:57:23 Integrated Voltage Measurement dXYus = -1.0001 mm, dXYds = 0.0000 mm dXY = -1.0001 mm VT = -89.461 uVs 14:57:23 Integrated Voltage Measurement... 14:57:32 Voltage Offset Correction: v_off_ini = 868.856 uV, v_off_fin = 881.183 uV 14:57:32 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 85.266 uVs 14:57:32 Integrated Voltage Measurement... 14:57:41 Voltage Offset Correction: v_off_ini = 895.747 uV, v_off_fin = 888.018 uV 14:57:41 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -80.460 uVs 14:57:41 Integrated Voltage Measurement... 14:57:50 Voltage Offset Correction: v_off_ini = 886.872 uV, v_off_fin = 891.769 uV 14:57:50 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 62.345 uVs 14:57:50 Integrated Voltage Measurement... 14:57:59 Voltage Offset Correction: v_off_ini = 890.298 uV, v_off_fin = 888.353 uV 14:57:59 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -71.786 uVs 14:57:59 Integrated Voltage Measurement... 14:58:08 Voltage Offset Correction: v_off_ini = 887.346 uV, v_off_fin = 899.183 uV 14:58:08 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 94.325 uVs 14:58:11 I2X Average and RMS of Measurements VT = -80.607 +- 6.742 uVs I2X = 7.416 +- 0.620 uTm^2 14:58:16 Wire Position: X0 = -1.0000 mm, Y0 = 0.0000 mm 14:58:18 Integrated Voltage Measurement... 14:58:27 Voltage Offset Correction: v_off_ini = 903.214 uV, v_off_fin = 884.103 uV 14:58:27 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0001 mm dXY = 1.0000 mm VT = -344.792 uVs 14:58:27 Integrated Voltage Measurement... 14:58:36 Voltage Offset Correction: v_off_ini = 905.707 uV, v_off_fin = 897.680 uV 14:58:36 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 317.015 uVs 14:58:36 Integrated Voltage Measurement... 14:58:45 Voltage Offset Correction: v_off_ini = 888.188 uV, v_off_fin = 904.674 uV 14:58:45 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -323.586 uVs 14:58:46 Integrated Voltage Measurement... 14:58:54 Voltage Offset Correction: v_off_ini = 896.577 uV, v_off_fin = 892.842 uV 14:58:55 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 320.954 uVs 14:58:55 Integrated Voltage Measurement... 14:59:04 Voltage Offset Correction: v_off_ini = 891.062 uV, v_off_fin = 903.183 uV 14:59:04 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -340.395 uVs 14:59:04 Integrated Voltage Measurement... 14:59:13 Voltage Offset Correction: v_off_ini = 911.408 uV, v_off_fin = 910.419 uV 14:59:13 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 333.785 uVs 14:59:15 I2Y Average and RMS of Measurements VT = -330.088 +- 6.078 uVs I2Y = -30.368 +- 0.559 uTm^2 14:59:20 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 14:59:24 Integrated Voltage Measurement... 14:59:33 Voltage Offset Correction: v_off_ini = 929.643 uV, v_off_fin = 929.973 uV 14:59:33 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -101.154 uVs 14:59:33 Integrated Voltage Measurement... 14:59:42 Voltage Offset Correction: v_off_ini = 904.520 uV, v_off_fin = 932.980 uV 14:59:42 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 78.957 uVs 14:59:42 Integrated Voltage Measurement... 14:59:51 Voltage Offset Correction: v_off_ini = 926.152 uV, v_off_fin = 941.675 uV 14:59:51 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -43.457 uVs 14:59:51 Integrated Voltage Measurement... 15:00:00 Voltage Offset Correction: v_off_ini = 940.501 uV, v_off_fin = 938.204 uV 15:00:00 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 37.555 uVs 15:00:00 Integrated Voltage Measurement... 15:00:09 Voltage Offset Correction: v_off_ini = 943.898 uV, v_off_fin = 955.090 uV 15:00:09 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -58.568 uVs 15:00:09 Integrated Voltage Measurement... 15:00:18 Voltage Offset Correction: v_off_ini = 920.102 uV, v_off_fin = 947.859 uV 15:00:18 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 78.021 uVs 15:00:22 I1X Average and RMS of Measurements VT = -66.285 +- 20.278 uVs I1X = 13.257 +- 4.056 uTm 15:00:25 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 15:00:29 Integrated Voltage Measurement... 15:00:38 Voltage Offset Correction: v_off_ini = 946.967 uV, v_off_fin = 943.015 uV 15:00:38 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -262.803 uVs 15:00:38 Integrated Voltage Measurement... 15:00:47 Voltage Offset Correction: v_off_ini = 943.287 uV, v_off_fin = 952.327 uV 15:00:47 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 255.564 uVs 15:00:47 Integrated Voltage Measurement... 15:00:56 Voltage Offset Correction: v_off_ini = 959.107 uV, v_off_fin = 938.162 uV 15:00:56 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -246.908 uVs 15:00:56 Integrated Voltage Measurement... 15:01:05 Voltage Offset Correction: v_off_ini = 928.771 uV, v_off_fin = 960.840 uV 15:01:05 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 250.315 uVs 15:01:05 Integrated Voltage Measurement... 15:01:14 Voltage Offset Correction: v_off_ini = 952.099 uV, v_off_fin = 935.342 uV 15:01:14 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -242.805 uVs 15:01:14 Integrated Voltage Measurement... 15:01:23 Voltage Offset Correction: v_off_ini = 939.506 uV, v_off_fin = 949.541 uV 15:01:23 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 262.108 uVs 15:01:27 I1Y Average and RMS of Measurements VT = -253.417 +- 4.369 uVs I1Y = -50.684 +- 0.874 uTm 15:01:32 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 15:01:35 Integrated Voltage Measurement... 15:01:44 Voltage Offset Correction: v_off_ini = 923.200 uV, v_off_fin = 919.626 uV 15:01:44 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -80.859 uVs 15:01:44 Integrated Voltage Measurement... 15:01:53 Voltage Offset Correction: v_off_ini = 926.482 uV, v_off_fin = 928.974 uV 15:01:53 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 55.771 uVs 15:01:53 Integrated Voltage Measurement... 15:02:02 Voltage Offset Correction: v_off_ini = 911.429 uV, v_off_fin = 908.083 uV 15:02:02 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -61.308 uVs 15:02:02 Integrated Voltage Measurement... 15:02:11 Voltage Offset Correction: v_off_ini = 932.915 uV, v_off_fin = 950.299 uV 15:02:11 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 66.819 uVs 15:02:11 Integrated Voltage Measurement... 15:02:20 Voltage Offset Correction: v_off_ini = 935.059 uV, v_off_fin = 931.036 uV 15:02:20 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -63.353 uVs 15:02:20 Integrated Voltage Measurement... 15:02:29 Voltage Offset Correction: v_off_ini = 936.141 uV, v_off_fin = 943.153 uV 15:02:29 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 76.117 uVs 15:02:32 I2X Average and RMS of Measurements VT = -67.371 +- 2.410 uVs I2X = 6.198 +- 0.222 uTm^2 15:02:37 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 15:02:39 Integrated Voltage Measurement... 15:02:48 Voltage Offset Correction: v_off_ini = 939.333 uV, v_off_fin = 917.826 uV 15:02:48 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -313.319 uVs 15:02:48 Integrated Voltage Measurement... 15:02:57 Voltage Offset Correction: v_off_ini = 937.695 uV, v_off_fin = 950.484 uV 15:02:57 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 299.771 uVs 15:02:57 Integrated Voltage Measurement... 15:03:06 Voltage Offset Correction: v_off_ini = 930.879 uV, v_off_fin = 927.778 uV 15:03:06 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -299.048 uVs 15:03:06 Integrated Voltage Measurement... 15:03:15 Voltage Offset Correction: v_off_ini = 912.582 uV, v_off_fin = 917.200 uV 15:03:15 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 338.973 uVs 15:03:15 Integrated Voltage Measurement... 15:03:24 Voltage Offset Correction: v_off_ini = 924.281 uV, v_off_fin = 934.264 uV 15:03:24 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -301.650 uVs 15:03:24 Integrated Voltage Measurement... 15:03:33 Voltage Offset Correction: v_off_ini = 933.963 uV, v_off_fin = 923.254 uV 15:03:33 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 350.633 uVs 15:03:36 I2Y Average and RMS of Measurements VT = -317.232 +- 8.098 uVs I2Y = -29.185 +- 0.745 uTm^2 15:03:41 Wire Position: X0 = 1.0000 mm, Y0 = 0.0000 mm 15:03:45 Integrated Voltage Measurement... 15:03:54 Voltage Offset Correction: v_off_ini = 921.616 uV, v_off_fin = 926.196 uV 15:03:54 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -63.769 uVs 15:03:54 Integrated Voltage Measurement... 15:04:03 Voltage Offset Correction: v_off_ini = 936.216 uV, v_off_fin = 915.924 uV 15:04:03 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 65.652 uVs 15:04:03 Integrated Voltage Measurement... 15:04:12 Voltage Offset Correction: v_off_ini = 919.841 uV, v_off_fin = 917.327 uV 15:04:12 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -55.719 uVs 15:04:12 Integrated Voltage Measurement... 15:04:21 Voltage Offset Correction: v_off_ini = 913.600 uV, v_off_fin = 914.035 uV 15:04:21 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 66.508 uVs 15:04:21 Integrated Voltage Measurement... 15:04:30 Voltage Offset Correction: v_off_ini = 917.054 uV, v_off_fin = 925.137 uV 15:04:30 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -65.679 uVs 15:04:30 Integrated Voltage Measurement... 15:04:39 Voltage Offset Correction: v_off_ini = 910.373 uV, v_off_fin = 927.203 uV 15:04:39 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 59.116 uVs 15:04:43 I1X Average and RMS of Measurements VT = -62.741 +- 1.488 uVs I1X = 12.548 +- 0.298 uTm 15:04:46 Wire Position: X0 = 1.0000 mm, Y0 = 0.0000 mm 15:04:50 Integrated Voltage Measurement... 15:04:59 Voltage Offset Correction: v_off_ini = 925.971 uV, v_off_fin = 918.105 uV 15:04:59 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -240.872 uVs 15:04:59 Integrated Voltage Measurement... 15:05:08 Voltage Offset Correction: v_off_ini = 932.052 uV, v_off_fin = 935.782 uV 15:05:08 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 242.853 uVs 15:05:08 Integrated Voltage Measurement... 15:05:17 Voltage Offset Correction: v_off_ini = 939.599 uV, v_off_fin = 907.350 uV 15:05:17 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -250.223 uVs 15:05:17 Integrated Voltage Measurement... 15:05:26 Voltage Offset Correction: v_off_ini = 926.344 uV, v_off_fin = 939.131 uV 15:05:26 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 262.893 uVs 15:05:26 Integrated Voltage Measurement... 15:05:35 Voltage Offset Correction: v_off_ini = 937.163 uV, v_off_fin = 922.775 uV 15:05:35 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -269.918 uVs 15:05:35 Integrated Voltage Measurement... 15:05:44 Voltage Offset Correction: v_off_ini = 927.469 uV, v_off_fin = 928.266 uV 15:05:44 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 260.110 uVs 15:05:48 I1Y Average and RMS of Measurements VT = -254.479 +- 9.565 uVs I1Y = -50.896 +- 1.913 uTm 15:05:53 Wire Position: X0 = 1.0000 mm, Y0 = 0.0000 mm 15:05:56 Integrated Voltage Measurement... 15:06:05 Voltage Offset Correction: v_off_ini = 912.083 uV, v_off_fin = 909.006 uV 15:06:05 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -69.788 uVs 15:06:05 Integrated Voltage Measurement... 15:06:14 Voltage Offset Correction: v_off_ini = 918.362 uV, v_off_fin = 929.473 uV 15:06:14 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 85.351 uVs 15:06:14 Integrated Voltage Measurement... 15:06:23 Voltage Offset Correction: v_off_ini = 934.981 uV, v_off_fin = 926.648 uV 15:06:23 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -68.086 uVs 15:06:23 Integrated Voltage Measurement... 15:06:32 Voltage Offset Correction: v_off_ini = 919.639 uV, v_off_fin = 922.352 uV 15:06:32 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 64.600 uVs 15:06:32 Integrated Voltage Measurement... 15:06:41 Voltage Offset Correction: v_off_ini = 917.075 uV, v_off_fin = 930.545 uV 15:06:41 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -74.953 uVs 15:06:41 Integrated Voltage Measurement... 15:06:50 Voltage Offset Correction: v_off_ini = 939.456 uV, v_off_fin = 957.608 uV 15:06:50 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 69.694 uVs 15:06:53 I2X Average and RMS of Measurements VT = -72.079 +- 4.586 uVs I2X = 6.631 +- 0.422 uTm^2 15:06:58 Wire Position: X0 = 1.0000 mm, Y0 = 0.0000 mm 15:07:00 Integrated Voltage Measurement... 15:07:09 Voltage Offset Correction: v_off_ini = 927.175 uV, v_off_fin = 952.179 uV 15:07:09 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -323.720 uVs 15:07:09 Integrated Voltage Measurement... 15:07:18 Voltage Offset Correction: v_off_ini = 959.154 uV, v_off_fin = 947.020 uV 15:07:18 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 340.045 uVs 15:07:18 Integrated Voltage Measurement... 15:07:27 Voltage Offset Correction: v_off_ini = 952.945 uV, v_off_fin = 960.011 uV 15:07:27 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -325.946 uVs 15:07:27 Integrated Voltage Measurement... 15:07:36 Voltage Offset Correction: v_off_ini = 942.521 uV, v_off_fin = 938.383 uV 15:07:36 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 318.510 uVs 15:07:36 Integrated Voltage Measurement... 15:07:45 Voltage Offset Correction: v_off_ini = 958.083 uV, v_off_fin = 946.236 uV 15:07:45 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -317.539 uVs 15:07:45 Integrated Voltage Measurement... 15:07:54 Voltage Offset Correction: v_off_ini = 952.873 uV, v_off_fin = 962.476 uV 15:07:54 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 327.048 uVs 15:07:57 I2Y Average and RMS of Measurements VT = -325.468 +- 4.536 uVs I2Y = -29.943 +- 0.417 uTm^2 15:08:01 Wire Position: X0 = 2.0000 mm, Y0 = 0.0000 mm 15:08:06 Integrated Voltage Measurement... 15:08:15 Voltage Offset Correction: v_off_ini = 970.318 uV, v_off_fin = 967.965 uV 15:08:15 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.4998 mm dXY = -0.4999 mm VT = -65.180 uVs 15:08:15 Integrated Voltage Measurement... 15:08:24 Voltage Offset Correction: v_off_ini = 974.079 uV, v_off_fin = 973.884 uV 15:08:24 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 67.899 uVs 15:08:24 Integrated Voltage Measurement... 15:08:33 Voltage Offset Correction: v_off_ini = 964.615 uV, v_off_fin = 972.664 uV 15:08:33 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -58.106 uVs 15:08:33 Integrated Voltage Measurement... 15:08:42 Voltage Offset Correction: v_off_ini = 990.355 uV, v_off_fin = 990.839 uV 15:08:42 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 44.419 uVs 15:08:42 Integrated Voltage Measurement... 15:08:51 Voltage Offset Correction: v_off_ini = 978.813 uV, v_off_fin = 979.239 uV 15:08:51 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -66.171 uVs 15:08:51 Integrated Voltage Measurement... 15:09:00 Voltage Offset Correction: v_off_ini = 976.310 uV, v_off_fin = 978.461 uV 15:09:00 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 68.062 uVs 15:09:04 I1X Average and RMS of Measurements VT = -61.639 +- 7.342 uVs I1X = 12.329 +- 1.469 uTm 15:09:07 Wire Position: X0 = 2.0000 mm, Y0 = 0.0000 mm 15:09:11 Integrated Voltage Measurement... 15:09:20 Voltage Offset Correction: v_off_ini = 979.854 uV, v_off_fin = 971.654 uV 15:09:20 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -241.267 uVs 15:09:20 Integrated Voltage Measurement... 15:09:29 Voltage Offset Correction: v_off_ini = 990.414 uV, v_off_fin = 1004.494 uV 15:09:29 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 245.803 uVs 15:09:29 Integrated Voltage Measurement... 15:09:38 Voltage Offset Correction: v_off_ini = 999.436 uV, v_off_fin = 979.337 uV 15:09:38 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -232.647 uVs 15:09:38 Integrated Voltage Measurement... 15:09:47 Voltage Offset Correction: v_off_ini = 980.210 uV, v_off_fin = 999.810 uV 15:09:47 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 254.472 uVs 15:09:47 Integrated Voltage Measurement... 15:09:56 Voltage Offset Correction: v_off_ini = 1007.749 uV, v_off_fin = 983.828 uV 15:09:56 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -248.133 uVs 15:09:56 Integrated Voltage Measurement... 15:10:05 Voltage Offset Correction: v_off_ini = 994.884 uV, v_off_fin = 1008.071 uV 15:10:05 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 250.422 uVs 15:10:09 I1Y Average and RMS of Measurements VT = -245.457 +- 2.701 uVs I1Y = -49.091 +- 0.540 uTm 15:10:14 Wire Position: X0 = 2.0000 mm, Y0 = 0.0000 mm 15:10:17 Integrated Voltage Measurement... 15:10:26 Voltage Offset Correction: v_off_ini = 996.372 uV, v_off_fin = 962.948 uV 15:10:26 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -57.136 uVs 15:10:26 Integrated Voltage Measurement... 15:10:35 Voltage Offset Correction: v_off_ini = 969.141 uV, v_off_fin = 997.406 uV 15:10:35 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 94.618 uVs 15:10:35 Integrated Voltage Measurement... 15:10:44 Voltage Offset Correction: v_off_ini = 995.044 uV, v_off_fin = 965.628 uV 15:10:44 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -89.795 uVs 15:10:44 Integrated Voltage Measurement... 15:10:53 Voltage Offset Correction: v_off_ini = 968.010 uV, v_off_fin = 974.654 uV 15:10:53 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 71.056 uVs 15:10:53 Integrated Voltage Measurement... 15:11:02 Voltage Offset Correction: v_off_ini = 972.400 uV, v_off_fin = 969.475 uV 15:11:02 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -72.394 uVs 15:11:02 Integrated Voltage Measurement... 15:11:11 Voltage Offset Correction: v_off_ini = 976.414 uV, v_off_fin = 981.363 uV 15:11:11 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 85.372 uVs 15:11:13 I2X Average and RMS of Measurements VT = -78.395 +- 1.889 uVs I2X = 7.212 +- 0.174 uTm^2 15:11:19 Wire Position: X0 = 2.0000 mm, Y0 = 0.0000 mm 15:11:21 Integrated Voltage Measurement... 15:11:30 Voltage Offset Correction: v_off_ini = 945.622 uV, v_off_fin = 964.128 uV 15:11:30 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -303.965 uVs 15:11:30 Integrated Voltage Measurement... 15:11:39 Voltage Offset Correction: v_off_ini = 979.900 uV, v_off_fin = 960.221 uV 15:11:39 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 296.780 uVs 15:11:39 Integrated Voltage Measurement... 15:11:48 Voltage Offset Correction: v_off_ini = 967.816 uV, v_off_fin = 968.314 uV 15:11:48 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -313.546 uVs 15:11:48 Integrated Voltage Measurement... 15:11:57 Voltage Offset Correction: v_off_ini = 982.175 uV, v_off_fin = 979.311 uV 15:11:57 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 316.101 uVs 15:11:57 Integrated Voltage Measurement... 15:12:06 Voltage Offset Correction: v_off_ini = 944.786 uV, v_off_fin = 975.070 uV 15:12:06 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -294.636 uVs 15:12:06 Integrated Voltage Measurement... 15:12:15 Voltage Offset Correction: v_off_ini = 970.195 uV, v_off_fin = 980.828 uV 15:12:15 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 330.443 uVs 15:12:18 I2Y Average and RMS of Measurements VT = -309.245 +- 6.343 uVs I2Y = -28.451 +- 0.584 uTm^2 15:12:26 Wire Position: X0 = 0.0000 mm, Y0 = -1.0000 mm 15:12:30 Integrated Voltage Measurement... 15:12:39 Voltage Offset Correction: v_off_ini = 948.340 uV, v_off_fin = 948.245 uV 15:12:39 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -71.250 uVs 15:12:39 Integrated Voltage Measurement... 15:12:48 Voltage Offset Correction: v_off_ini = 953.490 uV, v_off_fin = 965.269 uV 15:12:48 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 79.293 uVs 15:12:48 Integrated Voltage Measurement... 15:12:57 Voltage Offset Correction: v_off_ini = 964.222 uV, v_off_fin = 978.696 uV 15:12:57 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -87.713 uVs 15:12:57 Integrated Voltage Measurement... 15:13:06 Voltage Offset Correction: v_off_ini = 976.584 uV, v_off_fin = 973.589 uV 15:13:06 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 71.798 uVs 15:13:06 Integrated Voltage Measurement... 15:13:15 Voltage Offset Correction: v_off_ini = 983.240 uV, v_off_fin = 992.609 uV 15:13:15 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -57.786 uVs 15:13:15 Integrated Voltage Measurement... 15:13:24 Voltage Offset Correction: v_off_ini = 988.775 uV, v_off_fin = 977.236 uV 15:13:24 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 62.016 uVs 15:13:28 I1X Average and RMS of Measurements VT = -71.643 +- 8.502 uVs I1X = 14.329 +- 1.700 uTm 15:13:31 Wire Position: X0 = 0.0000 mm, Y0 = -1.0000 mm 15:13:35 Integrated Voltage Measurement... 15:13:44 Voltage Offset Correction: v_off_ini = 962.911 uV, v_off_fin = 965.209 uV 15:13:44 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -268.451 uVs 15:13:44 Integrated Voltage Measurement... 15:13:53 Voltage Offset Correction: v_off_ini = 967.648 uV, v_off_fin = 969.048 uV 15:13:53 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 264.205 uVs 15:13:53 Integrated Voltage Measurement... 15:14:02 Voltage Offset Correction: v_off_ini = 955.494 uV, v_off_fin = 936.615 uV 15:14:02 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -271.730 uVs 15:14:02 Integrated Voltage Measurement... 15:14:11 Voltage Offset Correction: v_off_ini = 964.956 uV, v_off_fin = 982.419 uV 15:14:11 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 266.880 uVs 15:14:11 Integrated Voltage Measurement... 15:14:20 Voltage Offset Correction: v_off_ini = 966.049 uV, v_off_fin = 929.374 uV 15:14:20 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -263.336 uVs 15:14:20 Integrated Voltage Measurement... 15:14:29 Voltage Offset Correction: v_off_ini = 932.383 uV, v_off_fin = 959.197 uV 15:14:29 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 267.932 uVs 15:14:33 I1Y Average and RMS of Measurements VT = -267.089 +- 1.592 uVs I1Y = -53.418 +- 0.318 uTm 15:14:36 Wire Position: X0 = 0.0000 mm, Y0 = -1.0000 mm 15:14:38 Integrated Voltage Measurement... 15:14:47 Voltage Offset Correction: v_off_ini = 955.967 uV, v_off_fin = 949.618 uV 15:14:47 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -96.690 uVs 15:14:47 Integrated Voltage Measurement... 15:14:56 Voltage Offset Correction: v_off_ini = 948.091 uV, v_off_fin = 951.492 uV 15:14:56 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 116.770 uVs 15:14:56 Integrated Voltage Measurement... 15:15:05 Voltage Offset Correction: v_off_ini = 945.579 uV, v_off_fin = 928.077 uV 15:15:05 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -96.021 uVs 15:15:05 Integrated Voltage Measurement... 15:15:14 Voltage Offset Correction: v_off_ini = 930.056 uV, v_off_fin = 945.167 uV 15:15:14 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 91.590 uVs 15:15:14 Integrated Voltage Measurement... 15:15:23 Voltage Offset Correction: v_off_ini = 958.246 uV, v_off_fin = 933.521 uV 15:15:23 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -105.344 uVs 15:15:24 Integrated Voltage Measurement... 15:15:32 Voltage Offset Correction: v_off_ini = 940.519 uV, v_off_fin = 943.910 uV 15:15:33 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 106.337 uVs 15:15:35 I2X Average and RMS of Measurements VT = -102.125 +- 5.894 uVs I2X = 9.396 +- 0.542 uTm^2 15:15:37 Wire Position: X0 = 0.0000 mm, Y0 = -1.0000 mm 15:15:40 Integrated Voltage Measurement... 15:15:49 Voltage Offset Correction: v_off_ini = 932.333 uV, v_off_fin = 952.440 uV 15:15:49 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -314.626 uVs 15:15:49 Integrated Voltage Measurement... 15:15:58 Voltage Offset Correction: v_off_ini = 934.118 uV, v_off_fin = 933.025 uV 15:15:58 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 332.975 uVs 15:15:58 Integrated Voltage Measurement... 15:16:07 Voltage Offset Correction: v_off_ini = 944.579 uV, v_off_fin = 934.310 uV 15:16:07 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -308.582 uVs 15:16:07 Integrated Voltage Measurement... 15:16:16 Voltage Offset Correction: v_off_ini = 921.268 uV, v_off_fin = 932.614 uV 15:16:16 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 336.506 uVs 15:16:16 Integrated Voltage Measurement... 15:16:25 Voltage Offset Correction: v_off_ini = 933.909 uV, v_off_fin = 933.794 uV 15:16:25 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -339.070 uVs 15:16:25 Integrated Voltage Measurement... 15:16:34 Voltage Offset Correction: v_off_ini = 956.815 uV, v_off_fin = 941.379 uV 15:16:34 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 332.701 uVs 15:16:37 I2Y Average and RMS of Measurements VT = -327.410 +- 6.015 uVs I2Y = -30.122 +- 0.553 uTm^2 15:16:41 Wire Position: X0 = 0.0000 mm, Y0 = -0.5000 mm 15:16:45 Integrated Voltage Measurement... 15:16:54 Voltage Offset Correction: v_off_ini = 942.403 uV, v_off_fin = 939.626 uV 15:16:54 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -93.902 uVs 15:16:54 Integrated Voltage Measurement... 15:17:03 Voltage Offset Correction: v_off_ini = 956.039 uV, v_off_fin = 950.464 uV 15:17:03 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 57.054 uVs 15:17:03 Integrated Voltage Measurement... 15:17:12 Voltage Offset Correction: v_off_ini = 950.284 uV, v_off_fin = 947.636 uV 15:17:12 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -43.014 uVs 15:17:12 Integrated Voltage Measurement... 15:17:21 Voltage Offset Correction: v_off_ini = 937.619 uV, v_off_fin = 923.247 uV 15:17:21 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 77.084 uVs 15:17:21 Integrated Voltage Measurement... 15:17:30 Voltage Offset Correction: v_off_ini = 946.761 uV, v_off_fin = 943.525 uV 15:17:30 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -62.332 uVs 15:17:30 Integrated Voltage Measurement... 15:17:39 Voltage Offset Correction: v_off_ini = 947.695 uV, v_off_fin = 948.490 uV 15:17:39 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 80.641 uVs 15:17:43 I1X Average and RMS of Measurements VT = -69.005 +- 6.539 uVs I1X = 13.801 +- 1.308 uTm 15:17:46 Wire Position: X0 = 0.0000 mm, Y0 = -0.5000 mm 15:17:50 Integrated Voltage Measurement... 15:17:59 Voltage Offset Correction: v_off_ini = 954.302 uV, v_off_fin = 941.140 uV 15:17:59 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -252.840 uVs 15:17:59 Integrated Voltage Measurement... 15:18:08 Voltage Offset Correction: v_off_ini = 958.968 uV, v_off_fin = 956.972 uV 15:18:08 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 252.237 uVs 15:18:08 Integrated Voltage Measurement... 15:18:17 Voltage Offset Correction: v_off_ini = 949.465 uV, v_off_fin = 933.292 uV 15:18:17 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -248.394 uVs 15:18:17 Integrated Voltage Measurement... 15:18:26 Voltage Offset Correction: v_off_ini = 946.237 uV, v_off_fin = 957.770 uV 15:18:26 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 238.947 uVs 15:18:26 Integrated Voltage Measurement... 15:18:35 Voltage Offset Correction: v_off_ini = 947.269 uV, v_off_fin = 948.303 uV 15:18:35 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -232.102 uVs 15:18:35 Integrated Voltage Measurement... 15:18:44 Voltage Offset Correction: v_off_ini = 945.110 uV, v_off_fin = 950.476 uV 15:18:44 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 238.750 uVs 15:18:48 I1Y Average and RMS of Measurements VT = -243.878 +- 6.987 uVs I1Y = -48.777 +- 1.399 uTm 15:18:51 Wire Position: X0 = 0.0000 mm, Y0 = -0.5000 mm 15:18:54 Integrated Voltage Measurement... 15:19:03 Voltage Offset Correction: v_off_ini = 971.270 uV, v_off_fin = 933.251 uV 15:19:03 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -85.231 uVs 15:19:03 Integrated Voltage Measurement... 15:19:12 Voltage Offset Correction: v_off_ini = 958.641 uV, v_off_fin = 968.032 uV 15:19:12 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 92.377 uVs 15:19:12 Integrated Voltage Measurement... 15:19:21 Voltage Offset Correction: v_off_ini = 947.653 uV, v_off_fin = 949.245 uV 15:19:21 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -63.047 uVs 15:19:21 Integrated Voltage Measurement... 15:19:30 Voltage Offset Correction: v_off_ini = 954.646 uV, v_off_fin = 967.132 uV 15:19:30 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 89.015 uVs 15:19:30 Integrated Voltage Measurement... 15:19:39 Voltage Offset Correction: v_off_ini = 943.221 uV, v_off_fin = 920.775 uV 15:19:39 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -99.409 uVs 15:19:39 Integrated Voltage Measurement... 15:19:48 Voltage Offset Correction: v_off_ini = 939.349 uV, v_off_fin = 968.611 uV 15:19:48 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 79.184 uVs 15:19:50 I2X Average and RMS of Measurements VT = -84.710 +- 6.141 uVs I2X = 7.793 +- 0.565 uTm^2 15:19:53 Wire Position: X0 = 0.0000 mm, Y0 = -0.5000 mm 15:19:55 Integrated Voltage Measurement... 15:20:04 Voltage Offset Correction: v_off_ini = 962.742 uV, v_off_fin = 952.592 uV 15:20:04 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -316.033 uVs 15:20:04 Integrated Voltage Measurement... 15:20:13 Voltage Offset Correction: v_off_ini = 934.409 uV, v_off_fin = 950.751 uV 15:20:13 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 319.262 uVs 15:20:13 Integrated Voltage Measurement... 15:20:22 Voltage Offset Correction: v_off_ini = 964.243 uV, v_off_fin = 933.814 uV 15:20:22 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -342.004 uVs 15:20:22 Integrated Voltage Measurement... 15:20:31 Voltage Offset Correction: v_off_ini = 914.658 uV, v_off_fin = 931.882 uV 15:20:31 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 324.088 uVs 15:20:31 Integrated Voltage Measurement... 15:20:40 Voltage Offset Correction: v_off_ini = 928.924 uV, v_off_fin = 936.931 uV 15:20:41 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -320.343 uVs 15:20:41 Integrated Voltage Measurement... 15:20:50 Voltage Offset Correction: v_off_ini = 959.962 uV, v_off_fin = 962.657 uV 15:20:50 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 319.430 uVs 15:20:52 I2Y Average and RMS of Measurements VT = -323.526 +- 6.793 uVs I2Y = -29.764 +- 0.625 uTm^2 15:20:56 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 15:21:00 Integrated Voltage Measurement... 15:21:09 Voltage Offset Correction: v_off_ini = 959.197 uV, v_off_fin = 970.935 uV 15:21:09 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -56.924 uVs 15:21:09 Integrated Voltage Measurement... 15:21:18 Voltage Offset Correction: v_off_ini = 979.909 uV, v_off_fin = 991.563 uV 15:21:18 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 41.068 uVs 15:21:18 Integrated Voltage Measurement... 15:21:27 Voltage Offset Correction: v_off_ini = 980.602 uV, v_off_fin = 970.808 uV 15:21:27 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -34.112 uVs 15:21:27 Integrated Voltage Measurement... 15:21:36 Voltage Offset Correction: v_off_ini = 976.876 uV, v_off_fin = 986.232 uV 15:21:36 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 68.478 uVs 15:21:36 Integrated Voltage Measurement... 15:21:45 Voltage Offset Correction: v_off_ini = 971.348 uV, v_off_fin = 983.829 uV 15:21:45 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -68.036 uVs 15:21:45 Integrated Voltage Measurement... 15:21:54 Voltage Offset Correction: v_off_ini = 982.932 uV, v_off_fin = 991.078 uV 15:21:54 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 64.347 uVs 15:21:59 I1X Average and RMS of Measurements VT = -55.494 +- 7.622 uVs I1X = 11.099 +- 1.524 uTm 15:22:01 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 15:22:05 Integrated Voltage Measurement... 15:22:14 Voltage Offset Correction: v_off_ini = 964.290 uV, v_off_fin = 965.842 uV 15:22:14 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -237.888 uVs 15:22:14 Integrated Voltage Measurement... 15:22:23 Voltage Offset Correction: v_off_ini = 975.829 uV, v_off_fin = 983.318 uV 15:22:23 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 254.847 uVs 15:22:23 Integrated Voltage Measurement... 15:22:32 Voltage Offset Correction: v_off_ini = 984.956 uV, v_off_fin = 956.680 uV 15:22:32 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -251.697 uVs 15:22:32 Integrated Voltage Measurement... 15:22:41 Voltage Offset Correction: v_off_ini = 988.147 uV, v_off_fin = 983.568 uV 15:22:41 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 246.102 uVs 15:22:41 Integrated Voltage Measurement... 15:22:50 Voltage Offset Correction: v_off_ini = 972.162 uV, v_off_fin = 964.438 uV 15:22:50 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -237.563 uVs 15:22:50 Integrated Voltage Measurement... 15:22:59 Voltage Offset Correction: v_off_ini = 960.965 uV, v_off_fin = 972.331 uV 15:22:59 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 226.826 uVs 15:23:04 I1Y Average and RMS of Measurements VT = -242.487 +- 7.351 uVs I1Y = -48.497 +- 1.470 uTm 15:23:06 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 15:23:09 Integrated Voltage Measurement... 15:23:18 Voltage Offset Correction: v_off_ini = 954.443 uV, v_off_fin = 950.121 uV 15:23:18 Integrated Voltage Measurement dXYus = -1.0001 mm, dXYds = 0.0000 mm dXY = -1.0001 mm VT = -61.958 uVs 15:23:18 Integrated Voltage Measurement... 15:23:27 Voltage Offset Correction: v_off_ini = 951.339 uV, v_off_fin = 979.549 uV 15:23:27 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 74.172 uVs 15:23:27 Integrated Voltage Measurement... 15:23:36 Voltage Offset Correction: v_off_ini = 965.353 uV, v_off_fin = 961.021 uV 15:23:36 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -63.960 uVs 15:23:36 Integrated Voltage Measurement... 15:23:45 Voltage Offset Correction: v_off_ini = 957.081 uV, v_off_fin = 962.626 uV 15:23:45 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 94.786 uVs 15:23:45 Integrated Voltage Measurement... 15:23:54 Voltage Offset Correction: v_off_ini = 956.058 uV, v_off_fin = 962.357 uV 15:23:54 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -50.381 uVs 15:23:54 Integrated Voltage Measurement... 15:24:03 Voltage Offset Correction: v_off_ini = 965.038 uV, v_off_fin = 953.374 uV 15:24:03 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 64.148 uVs 15:24:05 I2X Average and RMS of Measurements VT = -68.234 +- 9.027 uVs I2X = 6.277 +- 0.830 uTm^2 15:24:08 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 15:24:10 Integrated Voltage Measurement... 15:24:19 Voltage Offset Correction: v_off_ini = 951.417 uV, v_off_fin = 959.992 uV 15:24:19 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -316.127 uVs 15:24:20 Integrated Voltage Measurement... 15:24:28 Voltage Offset Correction: v_off_ini = 951.264 uV, v_off_fin = 945.133 uV 15:24:29 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 316.625 uVs 15:24:29 Integrated Voltage Measurement... 15:24:38 Voltage Offset Correction: v_off_ini = 956.076 uV, v_off_fin = 953.646 uV 15:24:38 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -294.233 uVs 15:24:38 Integrated Voltage Measurement... 15:24:47 Voltage Offset Correction: v_off_ini = 950.874 uV, v_off_fin = 960.731 uV 15:24:47 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 338.525 uVs 15:24:47 Integrated Voltage Measurement... 15:24:56 Voltage Offset Correction: v_off_ini = 969.539 uV, v_off_fin = 955.937 uV 15:24:56 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -320.076 uVs 15:24:56 Integrated Voltage Measurement... 15:25:05 Voltage Offset Correction: v_off_ini = 951.071 uV, v_off_fin = 963.643 uV 15:25:05 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 333.733 uVs 15:25:07 I2Y Average and RMS of Measurements VT = -319.886 +- 4.963 uVs I2Y = -29.430 +- 0.457 uTm^2 15:25:11 Wire Position: X0 = 0.0000 mm, Y0 = 0.5000 mm 15:25:15 Integrated Voltage Measurement... 15:25:24 Voltage Offset Correction: v_off_ini = 964.351 uV, v_off_fin = 955.314 uV 15:25:24 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -55.767 uVs 15:25:24 Integrated Voltage Measurement... 15:25:33 Voltage Offset Correction: v_off_ini = 943.278 uV, v_off_fin = 940.293 uV 15:25:33 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 96.172 uVs 15:25:34 Integrated Voltage Measurement... 15:25:42 Voltage Offset Correction: v_off_ini = 954.517 uV, v_off_fin = 958.332 uV 15:25:43 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -47.888 uVs 15:25:43 Integrated Voltage Measurement... 15:25:52 Voltage Offset Correction: v_off_ini = 948.778 uV, v_off_fin = 940.620 uV 15:25:52 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 49.517 uVs 15:25:52 Integrated Voltage Measurement... 15:26:01 Voltage Offset Correction: v_off_ini = 935.559 uV, v_off_fin = 948.997 uV 15:26:01 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -54.914 uVs 15:26:01 Integrated Voltage Measurement... 15:26:10 Voltage Offset Correction: v_off_ini = 932.102 uV, v_off_fin = 927.714 uV 15:26:10 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 84.646 uVs 15:26:14 I1X Average and RMS of Measurements VT = -64.817 +- 11.672 uVs I1X = 12.964 +- 2.335 uTm 15:26:16 Wire Position: X0 = 0.0000 mm, Y0 = 0.5000 mm 15:26:21 Integrated Voltage Measurement... 15:26:30 Voltage Offset Correction: v_off_ini = 950.423 uV, v_off_fin = 948.722 uV 15:26:30 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = -247.786 uVs 15:26:30 Integrated Voltage Measurement... 15:26:39 Voltage Offset Correction: v_off_ini = 958.674 uV, v_off_fin = 946.846 uV 15:26:39 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 256.670 uVs 15:26:39 Integrated Voltage Measurement... 15:26:48 Voltage Offset Correction: v_off_ini = 933.163 uV, v_off_fin = 916.035 uV 15:26:48 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -265.836 uVs 15:26:48 Integrated Voltage Measurement... 15:26:57 Voltage Offset Correction: v_off_ini = 927.474 uV, v_off_fin = 939.786 uV 15:26:57 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5002 mm dXY = -0.5001 mm VT = 266.759 uVs 15:26:57 Integrated Voltage Measurement... 15:27:06 Voltage Offset Correction: v_off_ini = 939.886 uV, v_off_fin = 932.972 uV 15:27:06 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -266.732 uVs 15:27:06 Integrated Voltage Measurement... 15:27:15 Voltage Offset Correction: v_off_ini = 943.993 uV, v_off_fin = 958.760 uV 15:27:15 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 260.185 uVs 15:27:19 I1Y Average and RMS of Measurements VT = -260.661 +- 6.075 uVs I1Y = -52.128 +- 1.212 uTm 15:27:21 Wire Position: X0 = 0.0000 mm, Y0 = 0.5000 mm 15:27:24 Integrated Voltage Measurement... 15:27:33 Voltage Offset Correction: v_off_ini = 921.188 uV, v_off_fin = 922.439 uV 15:27:33 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -79.242 uVs 15:27:33 Integrated Voltage Measurement... 15:27:42 Voltage Offset Correction: v_off_ini = 920.478 uV, v_off_fin = 941.694 uV 15:27:42 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 80.094 uVs 15:27:42 Integrated Voltage Measurement... 15:27:51 Voltage Offset Correction: v_off_ini = 923.382 uV, v_off_fin = 930.895 uV 15:27:51 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -65.634 uVs 15:27:51 Integrated Voltage Measurement... 15:28:00 Voltage Offset Correction: v_off_ini = 942.630 uV, v_off_fin = 945.396 uV 15:28:00 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 60.224 uVs 15:28:00 Integrated Voltage Measurement... 15:28:09 Voltage Offset Correction: v_off_ini = 938.164 uV, v_off_fin = 940.878 uV 15:28:09 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -63.539 uVs 15:28:09 Integrated Voltage Measurement... 15:28:18 Voltage Offset Correction: v_off_ini = 942.165 uV, v_off_fin = 946.442 uV 15:28:18 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 52.943 uVs 15:28:21 I2X Average and RMS of Measurements VT = -66.946 +- 9.197 uVs I2X = 6.159 +- 0.846 uTm^2 15:28:23 Wire Position: X0 = 0.0000 mm, Y0 = 0.5000 mm 15:28:26 Integrated Voltage Measurement... 15:28:35 Voltage Offset Correction: v_off_ini = 930.784 uV, v_off_fin = 944.639 uV 15:28:35 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -337.081 uVs 15:28:35 Integrated Voltage Measurement... 15:28:44 Voltage Offset Correction: v_off_ini = 917.181 uV, v_off_fin = 925.879 uV 15:28:44 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 317.872 uVs 15:28:44 Integrated Voltage Measurement... 15:28:53 Voltage Offset Correction: v_off_ini = 914.860 uV, v_off_fin = 920.515 uV 15:28:53 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -305.888 uVs 15:28:53 Integrated Voltage Measurement... 15:29:02 Voltage Offset Correction: v_off_ini = 912.139 uV, v_off_fin = 928.574 uV 15:29:02 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 343.583 uVs 15:29:02 Integrated Voltage Measurement... 15:29:11 Voltage Offset Correction: v_off_ini = 914.812 uV, v_off_fin = 917.989 uV 15:29:11 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -330.844 uVs 15:29:11 Integrated Voltage Measurement... 15:29:20 Voltage Offset Correction: v_off_ini = 922.960 uV, v_off_fin = 941.057 uV 15:29:20 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 314.371 uVs 15:29:22 I2Y Average and RMS of Measurements VT = -324.940 +- 1.993 uVs I2Y = -29.894 +- 0.183 uTm^2 15:29:26 Wire Position: X0 = 0.0000 mm, Y0 = 1.0000 mm 15:29:31 Integrated Voltage Measurement... 15:29:40 Voltage Offset Correction: v_off_ini = 948.263 uV, v_off_fin = 937.677 uV 15:29:40 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -76.402 uVs 15:29:40 Integrated Voltage Measurement... 15:29:49 Voltage Offset Correction: v_off_ini = 928.627 uV, v_off_fin = 948.788 uV 15:29:49 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 39.520 uVs 15:29:49 Integrated Voltage Measurement... 15:29:58 Voltage Offset Correction: v_off_ini = 942.117 uV, v_off_fin = 954.678 uV 15:29:58 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -60.377 uVs 15:29:58 Integrated Voltage Measurement... 15:30:07 Voltage Offset Correction: v_off_ini = 937.758 uV, v_off_fin = 943.454 uV 15:30:07 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 57.513 uVs 15:30:07 Integrated Voltage Measurement... 15:30:16 Voltage Offset Correction: v_off_ini = 959.420 uV, v_off_fin = 949.484 uV 15:30:16 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -52.639 uVs 15:30:16 Integrated Voltage Measurement... 15:30:25 Voltage Offset Correction: v_off_ini = 940.016 uV, v_off_fin = 925.570 uV 15:30:25 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 71.634 uVs 15:30:29 I1X Average and RMS of Measurements VT = -59.681 +- 1.782 uVs I1X = 11.936 +- 0.356 uTm 15:30:31 Wire Position: X0 = 0.0000 mm, Y0 = 1.0000 mm 15:30:36 Integrated Voltage Measurement... 15:30:45 Voltage Offset Correction: v_off_ini = 923.086 uV, v_off_fin = 909.813 uV 15:30:45 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -249.372 uVs 15:30:45 Integrated Voltage Measurement... 15:30:54 Voltage Offset Correction: v_off_ini = 944.288 uV, v_off_fin = 956.740 uV 15:30:54 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 248.972 uVs 15:30:54 Integrated Voltage Measurement... 15:31:03 Voltage Offset Correction: v_off_ini = 946.019 uV, v_off_fin = 915.727 uV 15:31:03 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -260.343 uVs 15:31:03 Integrated Voltage Measurement... 15:31:12 Voltage Offset Correction: v_off_ini = 918.235 uV, v_off_fin = 931.810 uV 15:31:12 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 252.186 uVs 15:31:12 Integrated Voltage Measurement... 15:31:21 Voltage Offset Correction: v_off_ini = 924.073 uV, v_off_fin = 916.825 uV 15:31:21 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -257.940 uVs 15:31:21 Integrated Voltage Measurement... 15:31:30 Voltage Offset Correction: v_off_ini = 922.744 uV, v_off_fin = 937.252 uV 15:31:30 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 255.136 uVs 15:31:34 I1Y Average and RMS of Measurements VT = -253.991 +- 3.410 uVs I1Y = -50.795 +- 0.680 uTm 15:31:36 Wire Position: X0 = 0.0000 mm, Y0 = 1.0000 mm 15:31:39 Integrated Voltage Measurement... 15:31:48 Voltage Offset Correction: v_off_ini = 913.451 uV, v_off_fin = 910.748 uV 15:31:48 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -57.135 uVs 15:31:48 Integrated Voltage Measurement... 15:31:57 Voltage Offset Correction: v_off_ini = 907.786 uV, v_off_fin = 920.897 uV 15:31:57 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 89.734 uVs 15:31:57 Integrated Voltage Measurement... 15:32:06 Voltage Offset Correction: v_off_ini = 934.720 uV, v_off_fin = 906.391 uV 15:32:06 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -56.977 uVs 15:32:06 Integrated Voltage Measurement... 15:32:15 Voltage Offset Correction: v_off_ini = 929.747 uV, v_off_fin = 934.839 uV 15:32:15 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 62.713 uVs 15:32:15 Integrated Voltage Measurement... 15:32:24 Voltage Offset Correction: v_off_ini = 912.279 uV, v_off_fin = 905.538 uV 15:32:24 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -67.960 uVs 15:32:24 Integrated Voltage Measurement... 15:32:33 Voltage Offset Correction: v_off_ini = 925.836 uV, v_off_fin = 937.320 uV 15:32:33 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 65.362 uVs 15:32:36 I2X Average and RMS of Measurements VT = -66.647 +- 5.548 uVs I2X = 6.132 +- 0.510 uTm^2 15:32:38 Wire Position: X0 = 0.0000 mm, Y0 = 1.0000 mm 15:32:41 Integrated Voltage Measurement... 15:32:50 Voltage Offset Correction: v_off_ini = 920.521 uV, v_off_fin = 927.932 uV 15:32:50 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -322.452 uVs 15:32:50 Integrated Voltage Measurement... 15:32:59 Voltage Offset Correction: v_off_ini = 914.329 uV, v_off_fin = 917.414 uV 15:32:59 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 313.920 uVs 15:32:59 Integrated Voltage Measurement... 15:33:08 Voltage Offset Correction: v_off_ini = 934.625 uV, v_off_fin = 932.874 uV 15:33:08 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -330.778 uVs 15:33:08 Integrated Voltage Measurement... 15:33:17 Voltage Offset Correction: v_off_ini = 928.313 uV, v_off_fin = 922.124 uV 15:33:17 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 282.495 uVs 15:33:17 Integrated Voltage Measurement... 15:33:26 Voltage Offset Correction: v_off_ini = 912.506 uV, v_off_fin = 913.731 uV 15:33:26 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -311.339 uVs 15:33:26 Integrated Voltage Measurement... 15:33:35 Voltage Offset Correction: v_off_ini = 919.394 uV, v_off_fin = 918.525 uV 15:33:35 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 331.662 uVs 15:33:37 I2Y Average and RMS of Measurements VT = -315.441 +- 6.371 uVs I2Y = -29.021 +- 0.586 uTm^2 15:33:53 Wire Position: X0 = -2.0000 mm, Y0 = 0.0000 mm 15:33:58 Integrated Voltage Measurement... 15:34:07 Voltage Offset Correction: v_off_ini = 940.684 uV, v_off_fin = 941.424 uV 15:34:07 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -57.316 uVs 15:34:07 Integrated Voltage Measurement... 15:34:16 Voltage Offset Correction: v_off_ini = 928.429 uV, v_off_fin = 941.718 uV 15:34:16 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 66.494 uVs 15:34:16 Integrated Voltage Measurement... 15:34:25 Voltage Offset Correction: v_off_ini = 936.032 uV, v_off_fin = 936.456 uV 15:34:25 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -38.821 uVs 15:34:25 Integrated Voltage Measurement... 15:34:34 Voltage Offset Correction: v_off_ini = 944.325 uV, v_off_fin = 953.133 uV 15:34:34 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 68.047 uVs 15:34:34 Integrated Voltage Measurement... 15:34:43 Voltage Offset Correction: v_off_ini = 950.086 uV, v_off_fin = 945.381 uV 15:34:43 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -54.100 uVs 15:34:43 Integrated Voltage Measurement... 15:34:52 Voltage Offset Correction: v_off_ini = 954.033 uV, v_off_fin = 952.881 uV 15:34:52 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 55.676 uVs 15:34:56 I1X Average and RMS of Measurements VT = -56.743 +- 3.698 uVs I1X = 11.349 +- 0.740 uTm 15:34:59 Wire Position: X0 = -2.0000 mm, Y0 = 0.0000 mm 15:35:03 Integrated Voltage Measurement... 15:35:12 Voltage Offset Correction: v_off_ini = 958.810 uV, v_off_fin = 945.005 uV 15:35:12 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -254.424 uVs 15:35:12 Integrated Voltage Measurement... 15:35:21 Voltage Offset Correction: v_off_ini = 955.826 uV, v_off_fin = 972.439 uV 15:35:21 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 234.426 uVs 15:35:21 Integrated Voltage Measurement... 15:35:30 Voltage Offset Correction: v_off_ini = 966.289 uV, v_off_fin = 958.160 uV 15:35:30 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -241.242 uVs 15:35:30 Integrated Voltage Measurement... 15:35:39 Voltage Offset Correction: v_off_ini = 959.741 uV, v_off_fin = 961.296 uV 15:35:39 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 250.861 uVs 15:35:39 Integrated Voltage Measurement... 15:35:48 Voltage Offset Correction: v_off_ini = 975.348 uV, v_off_fin = 938.959 uV 15:35:48 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -265.987 uVs 15:35:48 Integrated Voltage Measurement... 15:35:57 Voltage Offset Correction: v_off_ini = 934.075 uV, v_off_fin = 963.100 uV 15:35:57 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 271.706 uVs 15:36:01 I1Y Average and RMS of Measurements VT = -253.108 +- 11.149 uVs I1Y = -50.622 +- 2.230 uTm 15:36:06 Wire Position: X0 = -2.0000 mm, Y0 = 0.0000 mm 15:36:09 Integrated Voltage Measurement... 15:36:18 Voltage Offset Correction: v_off_ini = 958.073 uV, v_off_fin = 945.479 uV 15:36:18 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -69.340 uVs 15:36:18 Integrated Voltage Measurement... 15:36:27 Voltage Offset Correction: v_off_ini = 946.566 uV, v_off_fin = 945.892 uV 15:36:27 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 74.992 uVs 15:36:27 Integrated Voltage Measurement... 15:36:36 Voltage Offset Correction: v_off_ini = 952.828 uV, v_off_fin = 950.092 uV 15:36:36 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -87.126 uVs 15:36:36 Integrated Voltage Measurement... 15:36:45 Voltage Offset Correction: v_off_ini = 933.548 uV, v_off_fin = 960.166 uV 15:36:45 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 78.618 uVs 15:36:45 Integrated Voltage Measurement... 15:36:54 Voltage Offset Correction: v_off_ini = 947.745 uV, v_off_fin = 950.271 uV 15:36:54 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -85.938 uVs 15:36:54 Integrated Voltage Measurement... 15:37:03 Voltage Offset Correction: v_off_ini = 922.825 uV, v_off_fin = 949.520 uV 15:37:03 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 89.082 uVs 15:37:05 I2X Average and RMS of Measurements VT = -80.849 +- 6.425 uVs I2X = 7.438 +- 0.591 uTm^2 15:37:10 Wire Position: X0 = -2.0000 mm, Y0 = 0.0000 mm 15:37:13 Integrated Voltage Measurement... 15:37:22 Voltage Offset Correction: v_off_ini = 954.213 uV, v_off_fin = 945.936 uV 15:37:22 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = -336.946 uVs 15:37:22 Integrated Voltage Measurement... 15:37:31 Voltage Offset Correction: v_off_ini = 919.043 uV, v_off_fin = 927.137 uV 15:37:31 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 331.689 uVs 15:37:31 Integrated Voltage Measurement... 15:37:40 Voltage Offset Correction: v_off_ini = 939.546 uV, v_off_fin = 953.336 uV 15:37:40 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -330.788 uVs 15:37:40 Integrated Voltage Measurement... 15:37:49 Voltage Offset Correction: v_off_ini = 963.432 uV, v_off_fin = 969.427 uV 15:37:49 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 321.879 uVs 15:37:49 Integrated Voltage Measurement... 15:37:58 Voltage Offset Correction: v_off_ini = 951.493 uV, v_off_fin = 952.128 uV 15:37:58 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -319.095 uVs 15:37:58 Integrated Voltage Measurement... 15:38:07 Voltage Offset Correction: v_off_ini = 953.147 uV, v_off_fin = 957.699 uV 15:38:07 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 342.739 uVs 15:38:09 I2Y Average and RMS of Measurements VT = -330.522 +- 3.271 uVs I2Y = -30.409 +- 0.302 uTm^2 15:38:14 Wire Position: X0 = -1.0000 mm, Y0 = 0.0000 mm 15:38:18 Integrated Voltage Measurement... 15:38:27 Voltage Offset Correction: v_off_ini = 956.762 uV, v_off_fin = 953.726 uV 15:38:27 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -48.815 uVs 15:38:27 Integrated Voltage Measurement... 15:38:36 Voltage Offset Correction: v_off_ini = 948.533 uV, v_off_fin = 954.818 uV 15:38:36 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 70.982 uVs 15:38:36 Integrated Voltage Measurement... 15:38:45 Voltage Offset Correction: v_off_ini = 953.137 uV, v_off_fin = 964.492 uV 15:38:45 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -100.014 uVs 15:38:45 Integrated Voltage Measurement... 15:38:54 Voltage Offset Correction: v_off_ini = 951.118 uV, v_off_fin = 940.117 uV 15:38:54 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 91.286 uVs 15:38:54 Integrated Voltage Measurement... 15:39:03 Voltage Offset Correction: v_off_ini = 954.460 uV, v_off_fin = 955.534 uV 15:39:03 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -76.593 uVs 15:39:03 Integrated Voltage Measurement... 15:39:12 Voltage Offset Correction: v_off_ini = 943.677 uV, v_off_fin = 935.728 uV 15:39:12 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 55.630 uVs 15:39:16 I1X Average and RMS of Measurements VT = -73.887 +- 15.597 uVs I1X = 14.777 +- 3.119 uTm 15:39:19 Wire Position: X0 = -1.0000 mm, Y0 = 0.0000 mm 15:39:23 Integrated Voltage Measurement... 15:39:32 Voltage Offset Correction: v_off_ini = 961.347 uV, v_off_fin = 951.399 uV 15:39:32 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -258.011 uVs 15:39:32 Integrated Voltage Measurement... 15:39:41 Voltage Offset Correction: v_off_ini = 952.217 uV, v_off_fin = 962.082 uV 15:39:41 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 245.838 uVs 15:39:41 Integrated Voltage Measurement... 15:39:50 Voltage Offset Correction: v_off_ini = 948.474 uV, v_off_fin = 936.871 uV 15:39:50 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = -253.357 uVs 15:39:50 Integrated Voltage Measurement... 15:39:59 Voltage Offset Correction: v_off_ini = 969.218 uV, v_off_fin = 970.460 uV 15:39:59 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 264.697 uVs 15:39:59 Integrated Voltage Measurement... 15:40:08 Voltage Offset Correction: v_off_ini = 959.692 uV, v_off_fin = 980.137 uV 15:40:08 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -268.479 uVs 15:40:08 Integrated Voltage Measurement... 15:40:17 Voltage Offset Correction: v_off_ini = 988.448 uV, v_off_fin = 989.569 uV 15:40:17 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 247.244 uVs 15:40:21 I1Y Average and RMS of Measurements VT = -256.271 +- 3.110 uVs I1Y = -51.255 +- 0.623 uTm 15:40:27 Wire Position: X0 = -1.0000 mm, Y0 = 0.0000 mm 15:40:29 Integrated Voltage Measurement... 15:40:38 Voltage Offset Correction: v_off_ini = 940.176 uV, v_off_fin = 964.068 uV 15:40:38 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -76.716 uVs 15:40:38 Integrated Voltage Measurement... 15:40:47 Voltage Offset Correction: v_off_ini = 983.771 uV, v_off_fin = 999.391 uV 15:40:47 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 64.722 uVs 15:40:47 Integrated Voltage Measurement... 15:40:56 Voltage Offset Correction: v_off_ini = 978.831 uV, v_off_fin = 987.350 uV 15:40:56 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -72.080 uVs 15:40:56 Integrated Voltage Measurement... 15:41:05 Voltage Offset Correction: v_off_ini = 994.626 uV, v_off_fin = 1003.339 uV 15:41:05 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 71.828 uVs 15:41:05 Integrated Voltage Measurement... 15:41:14 Voltage Offset Correction: v_off_ini = 961.137 uV, v_off_fin = 955.092 uV 15:41:14 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -60.380 uVs 15:41:14 Integrated Voltage Measurement... 15:41:23 Voltage Offset Correction: v_off_ini = 963.541 uV, v_off_fin = 980.146 uV 15:41:23 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 64.814 uVs 15:41:26 I2X Average and RMS of Measurements VT = -68.423 +- 4.150 uVs I2X = 6.295 +- 0.382 uTm^2 15:41:31 Wire Position: X0 = -1.0000 mm, Y0 = 0.0000 mm 15:41:33 Integrated Voltage Measurement... 15:41:42 Voltage Offset Correction: v_off_ini = 977.786 uV, v_off_fin = 987.600 uV 15:41:42 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -311.833 uVs 15:41:42 Integrated Voltage Measurement... 15:41:51 Voltage Offset Correction: v_off_ini = 986.268 uV, v_off_fin = 979.835 uV 15:41:51 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 320.326 uVs 15:41:51 Integrated Voltage Measurement... 15:42:00 Voltage Offset Correction: v_off_ini = 977.003 uV, v_off_fin = 963.334 uV 15:42:00 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -319.599 uVs 15:42:00 Integrated Voltage Measurement... 15:42:09 Voltage Offset Correction: v_off_ini = 967.540 uV, v_off_fin = 982.880 uV 15:42:09 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 318.123 uVs 15:42:09 Integrated Voltage Measurement... 15:42:18 Voltage Offset Correction: v_off_ini = 958.663 uV, v_off_fin = 969.602 uV 15:42:18 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -312.158 uVs 15:42:18 Integrated Voltage Measurement... 15:42:27 Voltage Offset Correction: v_off_ini = 981.368 uV, v_off_fin = 969.647 uV 15:42:28 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 311.820 uVs 15:42:30 I2Y Average and RMS of Measurements VT = -315.643 +- 2.822 uVs I2Y = -29.039 +- 0.260 uTm^2 15:42:35 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 15:42:39 Integrated Voltage Measurement... 15:42:48 Voltage Offset Correction: v_off_ini = 961.861 uV, v_off_fin = 976.513 uV 15:42:48 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -64.022 uVs 15:42:48 Integrated Voltage Measurement... 15:42:57 Voltage Offset Correction: v_off_ini = 961.739 uV, v_off_fin = 953.128 uV 15:42:57 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 72.822 uVs 15:42:57 Integrated Voltage Measurement... 15:43:06 Voltage Offset Correction: v_off_ini = 974.722 uV, v_off_fin = 977.922 uV 15:43:06 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -70.058 uVs 15:43:06 Integrated Voltage Measurement... 15:43:15 Voltage Offset Correction: v_off_ini = 972.934 uV, v_off_fin = 971.033 uV 15:43:15 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 60.481 uVs 15:43:15 Integrated Voltage Measurement... 15:43:24 Voltage Offset Correction: v_off_ini = 969.757 uV, v_off_fin = 969.477 uV 15:43:24 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -58.588 uVs 15:43:24 Integrated Voltage Measurement... 15:43:33 Voltage Offset Correction: v_off_ini = 982.532 uV, v_off_fin = 984.007 uV 15:43:33 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 58.109 uVs 15:43:37 I1X Average and RMS of Measurements VT = -64.013 +- 4.207 uVs I1X = 12.803 +- 0.842 uTm 15:43:40 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 15:43:44 Integrated Voltage Measurement... 15:43:53 Voltage Offset Correction: v_off_ini = 974.714 uV, v_off_fin = 975.273 uV 15:43:53 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = -251.322 uVs 15:43:53 Integrated Voltage Measurement... 15:44:02 Voltage Offset Correction: v_off_ini = 997.607 uV, v_off_fin = 980.476 uV 15:44:02 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 242.161 uVs 15:44:02 Integrated Voltage Measurement... 15:44:11 Voltage Offset Correction: v_off_ini = 963.909 uV, v_off_fin = 970.468 uV 15:44:11 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -255.391 uVs 15:44:11 Integrated Voltage Measurement... 15:44:20 Voltage Offset Correction: v_off_ini = 979.471 uV, v_off_fin = 972.709 uV 15:44:20 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 229.086 uVs 15:44:20 Integrated Voltage Measurement... 15:44:29 Voltage Offset Correction: v_off_ini = 974.579 uV, v_off_fin = 964.788 uV 15:44:29 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -246.861 uVs 15:44:29 Integrated Voltage Measurement... 15:44:38 Voltage Offset Correction: v_off_ini = 978.170 uV, v_off_fin = 980.837 uV 15:44:38 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 243.685 uVs 15:44:42 I1Y Average and RMS of Measurements VT = -244.751 +- 1.875 uVs I1Y = -48.951 +- 0.376 uTm 15:44:47 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 15:44:50 Integrated Voltage Measurement... 15:44:59 Voltage Offset Correction: v_off_ini = 974.101 uV, v_off_fin = 966.634 uV 15:44:59 Integrated Voltage Measurement dXYus = -1.0001 mm, dXYds = 0.0000 mm dXY = -1.0001 mm VT = -74.343 uVs 15:44:59 Integrated Voltage Measurement... 15:45:08 Voltage Offset Correction: v_off_ini = 974.973 uV, v_off_fin = 987.688 uV 15:45:08 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 42.654 uVs 15:45:08 Integrated Voltage Measurement... 15:45:17 Voltage Offset Correction: v_off_ini = 977.358 uV, v_off_fin = 964.313 uV 15:45:17 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -59.040 uVs 15:45:17 Integrated Voltage Measurement... 15:45:26 Voltage Offset Correction: v_off_ini = 967.724 uV, v_off_fin = 970.871 uV 15:45:26 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 58.447 uVs 15:45:26 Integrated Voltage Measurement... 15:45:35 Voltage Offset Correction: v_off_ini = 950.892 uV, v_off_fin = 968.208 uV 15:45:35 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -89.358 uVs 15:45:35 Integrated Voltage Measurement... 15:45:44 Voltage Offset Correction: v_off_ini = 969.690 uV, v_off_fin = 998.987 uV 15:45:44 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 87.168 uVs 15:45:46 I2X Average and RMS of Measurements VT = -68.501 +- 13.974 uVs I2X = 6.302 +- 1.286 uTm^2 15:45:51 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 15:45:54 Integrated Voltage Measurement... 15:46:03 Voltage Offset Correction: v_off_ini = 976.918 uV, v_off_fin = 963.068 uV 15:46:03 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -318.073 uVs 15:46:03 Integrated Voltage Measurement... 15:46:12 Voltage Offset Correction: v_off_ini = 967.064 uV, v_off_fin = 982.782 uV 15:46:12 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 303.690 uVs 15:46:12 Integrated Voltage Measurement... 15:46:21 Voltage Offset Correction: v_off_ini = 980.346 uV, v_off_fin = 980.624 uV 15:46:21 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -325.480 uVs 15:46:21 Integrated Voltage Measurement... 15:46:30 Voltage Offset Correction: v_off_ini = 985.757 uV, v_off_fin = 989.187 uV 15:46:30 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 309.749 uVs 15:46:30 Integrated Voltage Measurement... 15:46:39 Voltage Offset Correction: v_off_ini = 990.643 uV, v_off_fin = 980.669 uV 15:46:39 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -300.307 uVs 15:46:39 Integrated Voltage Measurement... 15:46:48 Voltage Offset Correction: v_off_ini = 996.758 uV, v_off_fin = 989.804 uV 15:46:48 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 329.248 uVs 15:46:50 I2Y Average and RMS of Measurements VT = -314.425 +- 2.760 uVs I2Y = -28.927 +- 0.254 uTm^2 15:46:55 Wire Position: X0 = 1.0000 mm, Y0 = 0.0000 mm 15:46:59 Integrated Voltage Measurement... 15:47:08 Voltage Offset Correction: v_off_ini = 1002.320 uV, v_off_fin = 1003.903 uV 15:47:08 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -62.075 uVs 15:47:08 Integrated Voltage Measurement... 15:47:17 Voltage Offset Correction: v_off_ini = 997.911 uV, v_off_fin = 1010.068 uV 15:47:17 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 66.132 uVs 15:47:17 Integrated Voltage Measurement... 15:47:26 Voltage Offset Correction: v_off_ini = 986.206 uV, v_off_fin = 988.631 uV 15:47:26 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -44.043 uVs 15:47:26 Integrated Voltage Measurement... 15:47:35 Voltage Offset Correction: v_off_ini = 990.471 uV, v_off_fin = 997.351 uV 15:47:35 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 74.365 uVs 15:47:35 Integrated Voltage Measurement... 15:47:44 Voltage Offset Correction: v_off_ini = 991.445 uV, v_off_fin = 987.578 uV 15:47:44 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -63.474 uVs 15:47:44 Integrated Voltage Measurement... 15:47:53 Voltage Offset Correction: v_off_ini = 991.645 uV, v_off_fin = 982.867 uV 15:47:53 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 59.056 uVs 15:47:58 I1X Average and RMS of Measurements VT = -61.524 +- 2.009 uVs I1X = 12.305 +- 0.402 uTm 15:48:00 Wire Position: X0 = 1.0000 mm, Y0 = 0.0000 mm 15:48:04 Integrated Voltage Measurement... 15:48:13 Voltage Offset Correction: v_off_ini = 997.730 uV, v_off_fin = 1002.876 uV 15:48:13 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -262.190 uVs 15:48:13 Integrated Voltage Measurement... 15:48:22 Voltage Offset Correction: v_off_ini = 998.590 uV, v_off_fin = 992.630 uV 15:48:22 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 256.144 uVs 15:48:22 Integrated Voltage Measurement... 15:48:31 Voltage Offset Correction: v_off_ini = 984.177 uV, v_off_fin = 987.290 uV 15:48:31 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -240.929 uVs 15:48:31 Integrated Voltage Measurement... 15:48:40 Voltage Offset Correction: v_off_ini = 1010.465 uV, v_off_fin = 998.140 uV 15:48:40 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 252.862 uVs 15:48:40 Integrated Voltage Measurement... 15:48:49 Voltage Offset Correction: v_off_ini = 994.804 uV, v_off_fin = 973.459 uV 15:48:49 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -241.974 uVs 15:48:49 Integrated Voltage Measurement... 15:48:58 Voltage Offset Correction: v_off_ini = 998.504 uV, v_off_fin = 1003.129 uV 15:48:58 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 256.007 uVs 15:49:02 I1Y Average and RMS of Measurements VT = -251.684 +- 5.360 uVs I1Y = -50.337 +- 1.072 uTm 15:49:08 Wire Position: X0 = 1.0000 mm, Y0 = 0.0000 mm 15:49:10 Integrated Voltage Measurement... 15:49:19 Voltage Offset Correction: v_off_ini = 996.164 uV, v_off_fin = 985.681 uV 15:49:19 Integrated Voltage Measurement dXYus = -1.0001 mm, dXYds = 0.0000 mm dXY = -1.0001 mm VT = -88.551 uVs 15:49:19 Integrated Voltage Measurement... 15:49:28 Voltage Offset Correction: v_off_ini = 991.915 uV, v_off_fin = 1017.900 uV 15:49:28 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 64.003 uVs 15:49:28 Integrated Voltage Measurement... 15:49:37 Voltage Offset Correction: v_off_ini = 1011.358 uV, v_off_fin = 997.335 uV 15:49:37 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -85.514 uVs 15:49:37 Integrated Voltage Measurement... 15:49:46 Voltage Offset Correction: v_off_ini = 983.832 uV, v_off_fin = 1010.717 uV 15:49:46 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 86.829 uVs 15:49:46 Integrated Voltage Measurement... 15:49:55 Voltage Offset Correction: v_off_ini = 957.217 uV, v_off_fin = 981.645 uV 15:49:55 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -77.778 uVs 15:49:55 Integrated Voltage Measurement... 15:50:04 Voltage Offset Correction: v_off_ini = 983.818 uV, v_off_fin = 1004.292 uV 15:50:04 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 87.234 uVs 15:50:07 I2X Average and RMS of Measurements VT = -81.652 +- 4.084 uVs I2X = 7.512 +- 0.376 uTm^2 15:50:12 Wire Position: X0 = 1.0000 mm, Y0 = 0.0000 mm 15:50:15 Integrated Voltage Measurement... 15:50:24 Voltage Offset Correction: v_off_ini = 994.883 uV, v_off_fin = 987.660 uV 15:50:24 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -332.264 uVs 15:50:24 Integrated Voltage Measurement... 15:50:33 Voltage Offset Correction: v_off_ini = 997.905 uV, v_off_fin = 999.438 uV 15:50:33 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 305.341 uVs 15:50:33 Integrated Voltage Measurement... 15:50:42 Voltage Offset Correction: v_off_ini = 993.792 uV, v_off_fin = 996.063 uV 15:50:42 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -294.129 uVs 15:50:42 Integrated Voltage Measurement... 15:50:51 Voltage Offset Correction: v_off_ini = 979.123 uV, v_off_fin = 1010.903 uV 15:50:51 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 307.283 uVs 15:50:51 Integrated Voltage Measurement... 15:51:00 Voltage Offset Correction: v_off_ini = 993.417 uV, v_off_fin = 988.080 uV 15:51:00 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -326.481 uVs 15:51:00 Integrated Voltage Measurement... 15:51:09 Voltage Offset Correction: v_off_ini = 991.685 uV, v_off_fin = 993.106 uV 15:51:09 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 319.287 uVs 15:51:11 I2Y Average and RMS of Measurements VT = -314.131 +- 9.638 uVs I2Y = -28.900 +- 0.887 uTm^2 15:51:16 Wire Position: X0 = 2.0000 mm, Y0 = 0.0000 mm 15:51:20 Integrated Voltage Measurement... 15:51:29 Voltage Offset Correction: v_off_ini = 999.486 uV, v_off_fin = 1000.967 uV 15:51:29 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -61.501 uVs 15:51:29 Integrated Voltage Measurement... 15:51:38 Voltage Offset Correction: v_off_ini = 991.235 uV, v_off_fin = 980.708 uV 15:51:38 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 73.773 uVs 15:51:38 Integrated Voltage Measurement... 15:51:47 Voltage Offset Correction: v_off_ini = 998.465 uV, v_off_fin = 1001.981 uV 15:51:47 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -58.153 uVs 15:51:47 Integrated Voltage Measurement... 15:51:56 Voltage Offset Correction: v_off_ini = 1002.389 uV, v_off_fin = 1001.122 uV 15:51:56 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 43.771 uVs 15:51:56 Integrated Voltage Measurement... 15:52:05 Voltage Offset Correction: v_off_ini = 988.523 uV, v_off_fin = 1017.075 uV 15:52:05 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -34.759 uVs 15:52:05 Integrated Voltage Measurement... 15:52:14 Voltage Offset Correction: v_off_ini = 1010.825 uV, v_off_fin = 1017.004 uV 15:52:14 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 78.507 uVs 15:52:18 I1X Average and RMS of Measurements VT = -58.411 +- 6.923 uVs I1X = 11.682 +- 1.385 uTm 15:52:21 Wire Position: X0 = 2.0000 mm, Y0 = 0.0000 mm 15:52:25 Integrated Voltage Measurement... 15:52:34 Voltage Offset Correction: v_off_ini = 1026.013 uV, v_off_fin = 984.965 uV 15:52:34 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -216.837 uVs 15:52:34 Integrated Voltage Measurement... 15:52:43 Voltage Offset Correction: v_off_ini = 1006.760 uV, v_off_fin = 1019.558 uV 15:52:43 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 231.320 uVs 15:52:43 Integrated Voltage Measurement... 15:52:52 Voltage Offset Correction: v_off_ini = 990.964 uV, v_off_fin = 1004.633 uV 15:52:52 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -235.733 uVs 15:52:52 Integrated Voltage Measurement... 15:53:01 Voltage Offset Correction: v_off_ini = 1023.422 uV, v_off_fin = 1026.345 uV 15:53:01 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 231.442 uVs 15:53:01 Integrated Voltage Measurement... 15:53:10 Voltage Offset Correction: v_off_ini = 987.735 uV, v_off_fin = 962.489 uV 15:53:10 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -218.730 uVs 15:53:10 Integrated Voltage Measurement... 15:53:19 Voltage Offset Correction: v_off_ini = 998.546 uV, v_off_fin = 1013.383 uV 15:53:19 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 243.970 uVs 15:53:23 I1Y Average and RMS of Measurements VT = -229.672 +- 4.059 uVs I1Y = -45.934 +- 0.812 uTm 15:53:29 Wire Position: X0 = 2.0000 mm, Y0 = 0.0000 mm 15:53:31 Integrated Voltage Measurement... 15:53:40 Voltage Offset Correction: v_off_ini = 993.612 uV, v_off_fin = 988.265 uV 15:53:40 Integrated Voltage Measurement dXYus = -1.0001 mm, dXYds = 0.0000 mm dXY = -1.0001 mm VT = -89.703 uVs 15:53:40 Integrated Voltage Measurement... 15:53:49 Voltage Offset Correction: v_off_ini = 1020.287 uV, v_off_fin = 1029.521 uV 15:53:49 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 89.643 uVs 15:53:49 Integrated Voltage Measurement... 15:53:58 Voltage Offset Correction: v_off_ini = 1003.461 uV, v_off_fin = 983.343 uV 15:53:58 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -75.471 uVs 15:53:58 Integrated Voltage Measurement... 15:54:07 Voltage Offset Correction: v_off_ini = 1006.207 uV, v_off_fin = 1010.236 uV 15:54:07 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 68.910 uVs 15:54:07 Integrated Voltage Measurement... 15:54:16 Voltage Offset Correction: v_off_ini = 1005.193 uV, v_off_fin = 1001.042 uV 15:54:16 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -54.030 uVs 15:54:16 Integrated Voltage Measurement... 15:54:25 Voltage Offset Correction: v_off_ini = 1002.197 uV, v_off_fin = 1001.422 uV 15:54:25 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 81.631 uVs 15:54:28 I2X Average and RMS of Measurements VT = -76.565 +- 9.438 uVs I2X = 7.044 +- 0.868 uTm^2 15:54:33 Wire Position: X0 = 2.0000 mm, Y0 = 0.0000 mm 15:54:35 Integrated Voltage Measurement... 15:54:44 Voltage Offset Correction: v_off_ini = 1027.409 uV, v_off_fin = 989.565 uV 15:54:44 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -307.692 uVs 15:54:44 Integrated Voltage Measurement... 15:54:53 Voltage Offset Correction: v_off_ini = 998.588 uV, v_off_fin = 985.360 uV 15:54:53 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 334.402 uVs 15:54:53 Integrated Voltage Measurement... 15:55:02 Voltage Offset Correction: v_off_ini = 977.021 uV, v_off_fin = 987.575 uV 15:55:02 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -305.372 uVs 15:55:03 Integrated Voltage Measurement... 15:55:11 Voltage Offset Correction: v_off_ini = 1011.219 uV, v_off_fin = 998.602 uV 15:55:12 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 275.053 uVs 15:55:12 Integrated Voltage Measurement... 15:55:21 Voltage Offset Correction: v_off_ini = 1005.883 uV, v_off_fin = 991.283 uV 15:55:21 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -314.325 uVs 15:55:21 Integrated Voltage Measurement... 15:55:30 Voltage Offset Correction: v_off_ini = 980.784 uV, v_off_fin = 996.176 uV 15:55:30 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 310.550 uVs 15:55:32 I2Y Average and RMS of Measurements VT = -307.899 +- 12.991 uVs I2Y = -28.327 +- 1.195 uTm^2 15:55:40 Wire Position: X0 = 0.0000 mm, Y0 = -1.0000 mm 15:55:44 Integrated Voltage Measurement... 15:55:53 Voltage Offset Correction: v_off_ini = 1010.215 uV, v_off_fin = 1011.807 uV 15:55:53 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -65.896 uVs 15:55:53 Integrated Voltage Measurement... 15:56:02 Voltage Offset Correction: v_off_ini = 995.013 uV, v_off_fin = 991.060 uV 15:56:02 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 86.272 uVs 15:56:02 Integrated Voltage Measurement... 15:56:11 Voltage Offset Correction: v_off_ini = 1001.261 uV, v_off_fin = 995.102 uV 15:56:11 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -75.603 uVs 15:56:11 Integrated Voltage Measurement... 15:56:20 Voltage Offset Correction: v_off_ini = 1006.714 uV, v_off_fin = 1012.647 uV 15:56:20 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 64.657 uVs 15:56:20 Integrated Voltage Measurement... 15:56:29 Voltage Offset Correction: v_off_ini = 1011.125 uV, v_off_fin = 998.624 uV 15:56:29 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -105.411 uVs 15:56:29 Integrated Voltage Measurement... 15:56:38 Voltage Offset Correction: v_off_ini = 1009.987 uV, v_off_fin = 1009.819 uV 15:56:38 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 79.618 uVs 15:56:42 I1X Average and RMS of Measurements VT = -79.576 +- 9.466 uVs I1X = 15.915 +- 1.893 uTm 15:56:45 Wire Position: X0 = 0.0000 mm, Y0 = -1.0000 mm 15:56:49 Integrated Voltage Measurement... 15:56:58 Voltage Offset Correction: v_off_ini = 997.234 uV, v_off_fin = 994.050 uV 15:56:58 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -253.559 uVs 15:56:58 Integrated Voltage Measurement... 15:57:07 Voltage Offset Correction: v_off_ini = 1002.542 uV, v_off_fin = 1002.911 uV 15:57:07 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 273.988 uVs 15:57:07 Integrated Voltage Measurement... 15:57:16 Voltage Offset Correction: v_off_ini = 1002.775 uV, v_off_fin = 997.022 uV 15:57:16 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -255.220 uVs 15:57:16 Integrated Voltage Measurement... 15:57:25 Voltage Offset Correction: v_off_ini = 1016.052 uV, v_off_fin = 992.863 uV 15:57:25 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 243.303 uVs 15:57:25 Integrated Voltage Measurement... 15:57:34 Voltage Offset Correction: v_off_ini = 1003.610 uV, v_off_fin = 1009.907 uV 15:57:34 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -250.408 uVs 15:57:34 Integrated Voltage Measurement... 15:57:43 Voltage Offset Correction: v_off_ini = 1002.451 uV, v_off_fin = 1035.009 uV 15:57:43 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 254.879 uVs 15:57:47 I1Y Average and RMS of Measurements VT = -255.226 +- 6.200 uVs I1Y = -51.045 +- 1.240 uTm 15:57:50 Wire Position: X0 = 0.0000 mm, Y0 = -1.0000 mm 15:57:52 Integrated Voltage Measurement... 15:58:01 Voltage Offset Correction: v_off_ini = 1002.590 uV, v_off_fin = 973.916 uV 15:58:01 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -92.551 uVs 15:58:02 Integrated Voltage Measurement... 15:58:10 Voltage Offset Correction: v_off_ini = 995.879 uV, v_off_fin = 1006.364 uV 15:58:11 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 94.754 uVs 15:58:11 Integrated Voltage Measurement... 15:58:20 Voltage Offset Correction: v_off_ini = 1024.872 uV, v_off_fin = 989.319 uV 15:58:20 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -119.604 uVs 15:58:20 Integrated Voltage Measurement... 15:58:29 Voltage Offset Correction: v_off_ini = 998.941 uV, v_off_fin = 999.868 uV 15:58:29 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 126.847 uVs 15:58:29 Integrated Voltage Measurement... 15:58:38 Voltage Offset Correction: v_off_ini = 1004.173 uV, v_off_fin = 1007.673 uV 15:58:38 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -95.057 uVs 15:58:38 Integrated Voltage Measurement... 15:58:47 Voltage Offset Correction: v_off_ini = 1000.603 uV, v_off_fin = 1006.859 uV 15:58:47 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 92.797 uVs 15:58:49 I2X Average and RMS of Measurements VT = -103.602 +- 13.876 uVs I2X = 9.531 +- 1.277 uTm^2 15:58:52 Wire Position: X0 = 0.0000 mm, Y0 = -1.0000 mm 15:58:54 Integrated Voltage Measurement... 15:59:03 Voltage Offset Correction: v_off_ini = 995.289 uV, v_off_fin = 973.086 uV 15:59:03 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -331.595 uVs 15:59:03 Integrated Voltage Measurement... 15:59:12 Voltage Offset Correction: v_off_ini = 995.792 uV, v_off_fin = 1008.093 uV 15:59:12 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 323.648 uVs 15:59:12 Integrated Voltage Measurement... 15:59:21 Voltage Offset Correction: v_off_ini = 1004.807 uV, v_off_fin = 991.201 uV 15:59:21 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -321.122 uVs 15:59:21 Integrated Voltage Measurement... 15:59:30 Voltage Offset Correction: v_off_ini = 1001.956 uV, v_off_fin = 991.064 uV 15:59:30 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 316.127 uVs 15:59:30 Integrated Voltage Measurement... 15:59:39 Voltage Offset Correction: v_off_ini = 977.932 uV, v_off_fin = 990.366 uV 15:59:39 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -311.931 uVs 15:59:39 Integrated Voltage Measurement... 15:59:48 Voltage Offset Correction: v_off_ini = 999.332 uV, v_off_fin = 1005.589 uV 15:59:48 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 323.125 uVs 15:59:51 I2Y Average and RMS of Measurements VT = -321.258 +- 4.522 uVs I2Y = -29.556 +- 0.416 uTm^2 15:59:55 Wire Position: X0 = 0.0000 mm, Y0 = -0.5000 mm 15:59:59 Integrated Voltage Measurement... 16:00:08 Voltage Offset Correction: v_off_ini = 1005.147 uV, v_off_fin = 1004.981 uV 16:00:08 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -72.517 uVs 16:00:08 Integrated Voltage Measurement... 16:00:17 Voltage Offset Correction: v_off_ini = 1009.005 uV, v_off_fin = 999.342 uV 16:00:17 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 73.653 uVs 16:00:17 Integrated Voltage Measurement... 16:00:26 Voltage Offset Correction: v_off_ini = 982.678 uV, v_off_fin = 986.404 uV 16:00:26 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -44.322 uVs 16:00:26 Integrated Voltage Measurement... 16:00:35 Voltage Offset Correction: v_off_ini = 979.883 uV, v_off_fin = 982.326 uV 16:00:35 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 48.173 uVs 16:00:35 Integrated Voltage Measurement... 16:00:44 Voltage Offset Correction: v_off_ini = 988.015 uV, v_off_fin = 1006.409 uV 16:00:44 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -62.235 uVs 16:00:44 Integrated Voltage Measurement... 16:00:53 Voltage Offset Correction: v_off_ini = 992.068 uV, v_off_fin = 1005.337 uV 16:00:53 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 54.261 uVs 16:00:57 I1X Average and RMS of Measurements VT = -59.193 +- 10.977 uVs I1X = 11.839 +- 2.195 uTm 16:01:00 Wire Position: X0 = 0.0000 mm, Y0 = -0.5000 mm 16:01:04 Integrated Voltage Measurement... 16:01:13 Voltage Offset Correction: v_off_ini = 997.687 uV, v_off_fin = 989.831 uV 16:01:13 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -262.326 uVs 16:01:13 Integrated Voltage Measurement... 16:01:22 Voltage Offset Correction: v_off_ini = 986.602 uV, v_off_fin = 998.789 uV 16:01:22 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 255.899 uVs 16:01:22 Integrated Voltage Measurement... 16:01:31 Voltage Offset Correction: v_off_ini = 1001.014 uV, v_off_fin = 999.697 uV 16:01:31 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -235.582 uVs 16:01:31 Integrated Voltage Measurement... 16:01:40 Voltage Offset Correction: v_off_ini = 990.906 uV, v_off_fin = 1008.411 uV 16:01:40 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 250.283 uVs 16:01:40 Integrated Voltage Measurement... 16:01:49 Voltage Offset Correction: v_off_ini = 990.072 uV, v_off_fin = 1001.583 uV 16:01:49 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -239.532 uVs 16:01:49 Integrated Voltage Measurement... 16:01:58 Voltage Offset Correction: v_off_ini = 1018.355 uV, v_off_fin = 1010.964 uV 16:01:58 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 257.032 uVs 16:02:02 I1Y Average and RMS of Measurements VT = -250.109 +- 6.731 uVs I1Y = -50.022 +- 1.346 uTm 16:02:05 Wire Position: X0 = 0.0000 mm, Y0 = -0.5000 mm 16:02:08 Integrated Voltage Measurement... 16:02:17 Voltage Offset Correction: v_off_ini = 992.157 uV, v_off_fin = 991.547 uV 16:02:17 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -83.477 uVs 16:02:17 Integrated Voltage Measurement... 16:02:26 Voltage Offset Correction: v_off_ini = 1004.414 uV, v_off_fin = 996.154 uV 16:02:26 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 105.761 uVs 16:02:26 Integrated Voltage Measurement... 16:02:35 Voltage Offset Correction: v_off_ini = 980.745 uV, v_off_fin = 989.387 uV 16:02:35 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -84.095 uVs 16:02:35 Integrated Voltage Measurement... 16:02:44 Voltage Offset Correction: v_off_ini = 1011.275 uV, v_off_fin = 1021.582 uV 16:02:44 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 72.347 uVs 16:02:44 Integrated Voltage Measurement... 16:02:53 Voltage Offset Correction: v_off_ini = 1002.371 uV, v_off_fin = 1003.383 uV 16:02:53 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -96.061 uVs 16:02:53 Integrated Voltage Measurement... 16:03:02 Voltage Offset Correction: v_off_ini = 1017.904 uV, v_off_fin = 1014.718 uV 16:03:02 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 76.227 uVs 16:03:04 I2X Average and RMS of Measurements VT = -86.328 +- 6.696 uVs I2X = 7.942 +- 0.616 uTm^2 16:03:07 Wire Position: X0 = 0.0000 mm, Y0 = -0.5000 mm 16:03:09 Integrated Voltage Measurement... 16:03:18 Voltage Offset Correction: v_off_ini = 1010.211 uV, v_off_fin = 1015.644 uV 16:03:18 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -324.383 uVs 16:03:18 Integrated Voltage Measurement... 16:03:27 Voltage Offset Correction: v_off_ini = 1014.807 uV, v_off_fin = 996.236 uV 16:03:27 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 315.850 uVs 16:03:27 Integrated Voltage Measurement... 16:03:36 Voltage Offset Correction: v_off_ini = 1003.206 uV, v_off_fin = 1011.552 uV 16:03:36 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -304.311 uVs 16:03:36 Integrated Voltage Measurement... 16:03:45 Voltage Offset Correction: v_off_ini = 1018.110 uV, v_off_fin = 1003.009 uV 16:03:45 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 319.244 uVs 16:03:45 Integrated Voltage Measurement... 16:03:54 Voltage Offset Correction: v_off_ini = 991.945 uV, v_off_fin = 1033.928 uV 16:03:54 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -308.893 uVs 16:03:55 Integrated Voltage Measurement... 16:04:04 Voltage Offset Correction: v_off_ini = 1001.357 uV, v_off_fin = 1010.771 uV 16:04:04 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 318.053 uVs 16:04:06 I2Y Average and RMS of Measurements VT = -315.123 +- 3.599 uVs I2Y = -28.991 +- 0.331 uTm^2 16:04:10 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:04:14 Integrated Voltage Measurement... 16:04:23 Voltage Offset Correction: v_off_ini = 1047.454 uV, v_off_fin = 1041.414 uV 16:04:23 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -74.841 uVs 16:04:23 Integrated Voltage Measurement... 16:04:32 Voltage Offset Correction: v_off_ini = 994.472 uV, v_off_fin = 1003.822 uV 16:04:32 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 95.215 uVs 16:04:32 Integrated Voltage Measurement... 16:04:41 Voltage Offset Correction: v_off_ini = 1018.082 uV, v_off_fin = 1021.490 uV 16:04:41 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -60.151 uVs 16:04:41 Integrated Voltage Measurement... 16:04:50 Voltage Offset Correction: v_off_ini = 1011.731 uV, v_off_fin = 1029.062 uV 16:04:50 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 39.572 uVs 16:04:50 Integrated Voltage Measurement... 16:04:59 Voltage Offset Correction: v_off_ini = 1010.852 uV, v_off_fin = 1030.164 uV 16:04:59 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -79.585 uVs 16:04:59 Integrated Voltage Measurement... 16:05:08 Voltage Offset Correction: v_off_ini = 1034.057 uV, v_off_fin = 1033.312 uV 16:05:08 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 55.431 uVs 16:05:13 I1X Average and RMS of Measurements VT = -67.466 +- 14.357 uVs I1X = 13.493 +- 2.871 uTm 16:05:15 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:05:19 Integrated Voltage Measurement... 16:05:28 Voltage Offset Correction: v_off_ini = 1029.244 uV, v_off_fin = 1000.808 uV 16:05:28 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = -253.659 uVs 16:05:28 Integrated Voltage Measurement... 16:05:37 Voltage Offset Correction: v_off_ini = 1036.805 uV, v_off_fin = 1035.442 uV 16:05:37 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 253.216 uVs 16:05:37 Integrated Voltage Measurement... 16:05:46 Voltage Offset Correction: v_off_ini = 1057.650 uV, v_off_fin = 1028.571 uV 16:05:46 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -236.462 uVs 16:05:46 Integrated Voltage Measurement... 16:05:55 Voltage Offset Correction: v_off_ini = 1028.302 uV, v_off_fin = 1043.015 uV 16:05:55 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 224.097 uVs 16:05:55 Integrated Voltage Measurement... 16:06:04 Voltage Offset Correction: v_off_ini = 1017.006 uV, v_off_fin = 1009.715 uV 16:06:04 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -237.725 uVs 16:06:04 Integrated Voltage Measurement... 16:06:13 Voltage Offset Correction: v_off_ini = 1004.801 uV, v_off_fin = 1017.393 uV 16:06:13 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 235.084 uVs 16:06:18 I1Y Average and RMS of Measurements VT = -240.041 +- 9.798 uVs I1Y = -48.009 +- 1.961 uTm 16:06:20 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:06:23 Integrated Voltage Measurement... 16:06:32 Voltage Offset Correction: v_off_ini = 1009.805 uV, v_off_fin = 1007.344 uV 16:06:32 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -74.681 uVs 16:06:32 Integrated Voltage Measurement... 16:06:41 Voltage Offset Correction: v_off_ini = 1021.419 uV, v_off_fin = 999.092 uV 16:06:41 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 75.042 uVs 16:06:41 Integrated Voltage Measurement... 16:06:50 Voltage Offset Correction: v_off_ini = 1018.789 uV, v_off_fin = 998.894 uV 16:06:50 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -96.939 uVs 16:06:50 Integrated Voltage Measurement... 16:06:59 Voltage Offset Correction: v_off_ini = 1005.199 uV, v_off_fin = 1013.364 uV 16:06:59 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 58.948 uVs 16:06:59 Integrated Voltage Measurement... 16:07:08 Voltage Offset Correction: v_off_ini = 1013.072 uV, v_off_fin = 1008.126 uV 16:07:08 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -85.087 uVs 16:07:08 Integrated Voltage Measurement... 16:07:17 Voltage Offset Correction: v_off_ini = 1021.395 uV, v_off_fin = 1013.280 uV 16:07:17 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 73.145 uVs 16:07:19 I2X Average and RMS of Measurements VT = -77.307 +- 1.794 uVs I2X = 7.112 +- 0.165 uTm^2 16:07:22 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:07:24 Integrated Voltage Measurement... 16:07:33 Voltage Offset Correction: v_off_ini = 998.686 uV, v_off_fin = 1014.519 uV 16:07:33 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -282.049 uVs 16:07:33 Integrated Voltage Measurement... 16:07:42 Voltage Offset Correction: v_off_ini = 1006.142 uV, v_off_fin = 1010.802 uV 16:07:42 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = 346.235 uVs 16:07:43 Integrated Voltage Measurement... 16:07:51 Voltage Offset Correction: v_off_ini = 999.885 uV, v_off_fin = 993.411 uV 16:07:52 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -294.431 uVs 16:07:52 Integrated Voltage Measurement... 16:08:01 Voltage Offset Correction: v_off_ini = 999.481 uV, v_off_fin = 988.437 uV 16:08:01 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 317.913 uVs 16:08:01 Integrated Voltage Measurement... 16:08:10 Voltage Offset Correction: v_off_ini = 1003.744 uV, v_off_fin = 1005.471 uV 16:08:10 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -335.022 uVs 16:08:10 Integrated Voltage Measurement... 16:08:19 Voltage Offset Correction: v_off_ini = 1020.091 uV, v_off_fin = 1032.519 uV 16:08:19 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 294.763 uVs 16:08:21 I2Y Average and RMS of Measurements VT = -311.736 +- 3.946 uVs I2Y = -28.680 +- 0.363 uTm^2 16:08:25 Wire Position: X0 = 0.0000 mm, Y0 = 0.5000 mm 16:08:29 Integrated Voltage Measurement... 16:08:38 Voltage Offset Correction: v_off_ini = 1004.608 uV, v_off_fin = 1022.078 uV 16:08:38 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -42.538 uVs 16:08:38 Integrated Voltage Measurement... 16:08:47 Voltage Offset Correction: v_off_ini = 1017.949 uV, v_off_fin = 1007.936 uV 16:08:47 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 78.675 uVs 16:08:47 Integrated Voltage Measurement... 16:08:56 Voltage Offset Correction: v_off_ini = 1020.319 uV, v_off_fin = 1008.857 uV 16:08:56 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -36.863 uVs 16:08:56 Integrated Voltage Measurement... 16:09:05 Voltage Offset Correction: v_off_ini = 1036.318 uV, v_off_fin = 1032.376 uV 16:09:05 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 69.839 uVs 16:09:05 Integrated Voltage Measurement... 16:09:14 Voltage Offset Correction: v_off_ini = 1024.838 uV, v_off_fin = 1019.457 uV 16:09:14 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -88.630 uVs 16:09:15 Integrated Voltage Measurement... 16:09:23 Voltage Offset Correction: v_off_ini = 1027.629 uV, v_off_fin = 1037.392 uV 16:09:24 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 41.336 uVs 16:09:28 I1X Average and RMS of Measurements VT = -59.647 +- 4.797 uVs I1X = 11.930 +- 0.959 uTm 16:09:30 Wire Position: X0 = 0.0000 mm, Y0 = 0.5000 mm 16:09:34 Integrated Voltage Measurement... 16:09:43 Voltage Offset Correction: v_off_ini = 1027.033 uV, v_off_fin = 1000.395 uV 16:09:43 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = -233.808 uVs 16:09:43 Integrated Voltage Measurement... 16:09:52 Voltage Offset Correction: v_off_ini = 1023.308 uV, v_off_fin = 1044.875 uV 16:09:52 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 256.277 uVs 16:09:53 Integrated Voltage Measurement... 16:10:01 Voltage Offset Correction: v_off_ini = 1014.695 uV, v_off_fin = 1029.337 uV 16:10:01 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5002 mm dXY = 0.5001 mm VT = -248.766 uVs 16:10:02 Integrated Voltage Measurement... 16:10:10 Voltage Offset Correction: v_off_ini = 1018.776 uV, v_off_fin = 1032.873 uV 16:10:10 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 231.431 uVs 16:10:11 Integrated Voltage Measurement... 16:10:19 Voltage Offset Correction: v_off_ini = 1036.359 uV, v_off_fin = 1007.672 uV 16:10:20 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -240.492 uVs 16:10:20 Integrated Voltage Measurement... 16:10:28 Voltage Offset Correction: v_off_ini = 1014.777 uV, v_off_fin = 1037.610 uV 16:10:29 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 240.661 uVs 16:10:33 I1Y Average and RMS of Measurements VT = -241.906 +- 2.226 uVs I1Y = -48.379 +- 0.449 uTm 16:10:35 Wire Position: X0 = 0.0000 mm, Y0 = 0.5000 mm 16:10:38 Integrated Voltage Measurement... 16:10:47 Voltage Offset Correction: v_off_ini = 1021.914 uV, v_off_fin = 1033.998 uV 16:10:47 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -57.212 uVs 16:10:47 Integrated Voltage Measurement... 16:10:56 Voltage Offset Correction: v_off_ini = 1030.838 uV, v_off_fin = 1047.108 uV 16:10:56 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 81.348 uVs 16:10:56 Integrated Voltage Measurement... 16:11:05 Voltage Offset Correction: v_off_ini = 1016.078 uV, v_off_fin = 1008.948 uV 16:11:05 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -84.278 uVs 16:11:05 Integrated Voltage Measurement... 16:11:14 Voltage Offset Correction: v_off_ini = 1022.564 uV, v_off_fin = 1023.095 uV 16:11:14 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 66.337 uVs 16:11:14 Integrated Voltage Measurement... 16:11:23 Voltage Offset Correction: v_off_ini = 1049.766 uV, v_off_fin = 1029.327 uV 16:11:23 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -77.813 uVs 16:11:23 Integrated Voltage Measurement... 16:11:32 Voltage Offset Correction: v_off_ini = 1042.073 uV, v_off_fin = 1033.610 uV 16:11:32 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 81.294 uVs 16:11:34 I2X Average and RMS of Measurements VT = -74.713 +- 4.215 uVs I2X = 6.874 +- 0.388 uTm^2 16:11:37 Wire Position: X0 = 0.0000 mm, Y0 = 0.5000 mm 16:11:40 Integrated Voltage Measurement... 16:11:49 Voltage Offset Correction: v_off_ini = 1045.377 uV, v_off_fin = 1026.924 uV 16:11:49 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -315.053 uVs 16:11:49 Integrated Voltage Measurement... 16:11:58 Voltage Offset Correction: v_off_ini = 1027.385 uV, v_off_fin = 1016.840 uV 16:11:58 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 297.457 uVs 16:11:58 Integrated Voltage Measurement... 16:12:07 Voltage Offset Correction: v_off_ini = 1040.271 uV, v_off_fin = 1026.879 uV 16:12:07 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -335.731 uVs 16:12:07 Integrated Voltage Measurement... 16:12:16 Voltage Offset Correction: v_off_ini = 1020.527 uV, v_off_fin = 1018.427 uV 16:12:16 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 318.201 uVs 16:12:16 Integrated Voltage Measurement... 16:12:25 Voltage Offset Correction: v_off_ini = 1019.535 uV, v_off_fin = 1019.302 uV 16:12:25 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -318.221 uVs 16:12:25 Integrated Voltage Measurement... 16:12:34 Voltage Offset Correction: v_off_ini = 1025.853 uV, v_off_fin = 1023.994 uV 16:12:34 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 329.438 uVs 16:12:36 I2Y Average and RMS of Measurements VT = -319.017 +- 9.114 uVs I2Y = -29.350 +- 0.839 uTm^2 16:12:40 Wire Position: X0 = 0.0000 mm, Y0 = 1.0000 mm 16:12:45 Integrated Voltage Measurement... 16:12:53 Voltage Offset Correction: v_off_ini = 1036.388 uV, v_off_fin = 1017.313 uV 16:12:54 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -57.964 uVs 16:12:54 Integrated Voltage Measurement... 16:13:03 Voltage Offset Correction: v_off_ini = 1010.277 uV, v_off_fin = 1036.484 uV 16:13:03 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 37.251 uVs 16:13:03 Integrated Voltage Measurement... 16:13:12 Voltage Offset Correction: v_off_ini = 1025.942 uV, v_off_fin = 1028.630 uV 16:13:12 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -82.408 uVs 16:13:12 Integrated Voltage Measurement... 16:13:21 Voltage Offset Correction: v_off_ini = 1017.025 uV, v_off_fin = 1030.876 uV 16:13:21 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = 55.485 uVs 16:13:21 Integrated Voltage Measurement... 16:13:30 Voltage Offset Correction: v_off_ini = 1031.590 uV, v_off_fin = 1036.383 uV 16:13:30 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -46.221 uVs 16:13:30 Integrated Voltage Measurement... 16:13:39 Voltage Offset Correction: v_off_ini = 1032.055 uV, v_off_fin = 1031.163 uV 16:13:39 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 35.118 uVs 16:13:43 I1X Average and RMS of Measurements VT = -52.408 +- 12.033 uVs I1X = 10.482 +- 2.407 uTm 16:13:45 Wire Position: X0 = 0.0000 mm, Y0 = 1.0000 mm 16:13:50 Integrated Voltage Measurement... 16:13:59 Voltage Offset Correction: v_off_ini = 1035.890 uV, v_off_fin = 1024.128 uV 16:13:59 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = -236.533 uVs 16:13:59 Integrated Voltage Measurement... 16:14:08 Voltage Offset Correction: v_off_ini = 1027.375 uV, v_off_fin = 1052.093 uV 16:14:08 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 279.214 uVs 16:14:08 Integrated Voltage Measurement... 16:14:17 Voltage Offset Correction: v_off_ini = 1042.068 uV, v_off_fin = 1021.278 uV 16:14:17 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -253.422 uVs 16:14:17 Integrated Voltage Measurement... 16:14:26 Voltage Offset Correction: v_off_ini = 1019.684 uV, v_off_fin = 1037.703 uV 16:14:26 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 246.428 uVs 16:14:26 Integrated Voltage Measurement... 16:14:35 Voltage Offset Correction: v_off_ini = 1049.556 uV, v_off_fin = 1040.269 uV 16:14:35 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -242.151 uVs 16:14:35 Integrated Voltage Measurement... 16:14:44 Voltage Offset Correction: v_off_ini = 1046.667 uV, v_off_fin = 1052.630 uV 16:14:44 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 255.369 uVs 16:14:48 I1Y Average and RMS of Measurements VT = -252.186 +- 4.050 uVs I1Y = -50.436 +- 0.812 uTm 16:14:50 Wire Position: X0 = 0.0000 mm, Y0 = 1.0000 mm 16:14:53 Integrated Voltage Measurement... 16:15:02 Voltage Offset Correction: v_off_ini = 1043.297 uV, v_off_fin = 1036.644 uV 16:15:02 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -74.903 uVs 16:15:02 Integrated Voltage Measurement... 16:15:11 Voltage Offset Correction: v_off_ini = 1044.121 uV, v_off_fin = 1043.336 uV 16:15:11 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 60.280 uVs 16:15:11 Integrated Voltage Measurement... 16:15:20 Voltage Offset Correction: v_off_ini = 1032.986 uV, v_off_fin = 1034.314 uV 16:15:20 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -47.603 uVs 16:15:20 Integrated Voltage Measurement... 16:15:29 Voltage Offset Correction: v_off_ini = 1035.499 uV, v_off_fin = 1036.403 uV 16:15:29 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 61.686 uVs 16:15:29 Integrated Voltage Measurement... 16:15:38 Voltage Offset Correction: v_off_ini = 1016.435 uV, v_off_fin = 1003.254 uV 16:15:38 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -66.111 uVs 16:15:38 Integrated Voltage Measurement... 16:15:47 Voltage Offset Correction: v_off_ini = 1010.943 uV, v_off_fin = 1022.774 uV 16:15:47 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 66.800 uVs 16:15:50 I2X Average and RMS of Measurements VT = -62.897 +- 5.854 uVs I2X = 5.787 +- 0.539 uTm^2 16:15:52 Wire Position: X0 = 0.0000 mm, Y0 = 1.0000 mm 16:15:55 Integrated Voltage Measurement... 16:16:04 Voltage Offset Correction: v_off_ini = 1039.464 uV, v_off_fin = 1039.477 uV 16:16:04 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -308.278 uVs 16:16:04 Integrated Voltage Measurement... 16:16:13 Voltage Offset Correction: v_off_ini = 1057.298 uV, v_off_fin = 1052.597 uV 16:16:13 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 321.010 uVs 16:16:13 Integrated Voltage Measurement... 16:16:22 Voltage Offset Correction: v_off_ini = 1050.154 uV, v_off_fin = 1061.286 uV 16:16:22 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -318.067 uVs 16:16:22 Integrated Voltage Measurement... 16:16:31 Voltage Offset Correction: v_off_ini = 1064.127 uV, v_off_fin = 1049.542 uV 16:16:31 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 324.264 uVs 16:16:31 Integrated Voltage Measurement... 16:16:40 Voltage Offset Correction: v_off_ini = 1047.009 uV, v_off_fin = 1048.312 uV 16:16:40 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -298.947 uVs 16:16:40 Integrated Voltage Measurement... 16:16:49 Voltage Offset Correction: v_off_ini = 1064.273 uV, v_off_fin = 1063.733 uV 16:16:49 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 315.109 uVs 16:16:51 I2Y Average and RMS of Measurements VT = -314.279 +- 5.777 uVs I2Y = -28.914 +- 0.532 uTm^2 16:17:21 Wire Position: X0 = -2.0000 mm, Y0 = 0.0000 mm 16:17:26 Integrated Voltage Measurement... 16:17:34 Voltage Offset Correction: v_off_ini = 1065.513 uV, v_off_fin = 1047.838 uV 16:17:35 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.4998 mm dXY = -0.4999 mm VT = -52.079 uVs 16:17:35 Integrated Voltage Measurement... 16:17:44 Voltage Offset Correction: v_off_ini = 1047.966 uV, v_off_fin = 1062.995 uV 16:17:44 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 63.808 uVs 16:17:44 Integrated Voltage Measurement... 16:17:53 Voltage Offset Correction: v_off_ini = 1072.287 uV, v_off_fin = 1084.539 uV 16:17:53 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -67.749 uVs 16:17:53 Integrated Voltage Measurement... 16:18:02 Voltage Offset Correction: v_off_ini = 1063.922 uV, v_off_fin = 1064.350 uV 16:18:02 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 93.236 uVs 16:18:02 Integrated Voltage Measurement... 16:18:11 Voltage Offset Correction: v_off_ini = 1087.689 uV, v_off_fin = 1063.078 uV 16:18:11 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -49.858 uVs 16:18:11 Integrated Voltage Measurement... 16:18:20 Voltage Offset Correction: v_off_ini = 1081.553 uV, v_off_fin = 1070.400 uV 16:18:20 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 51.820 uVs 16:18:24 I1X Average and RMS of Measurements VT = -63.092 +- 12.642 uVs I1X = 12.619 +- 2.528 uTm 16:18:26 Wire Position: X0 = -2.0000 mm, Y0 = 0.0000 mm 16:18:31 Integrated Voltage Measurement... 16:18:40 Voltage Offset Correction: v_off_ini = 1088.489 uV, v_off_fin = 1077.286 uV 16:18:40 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -260.011 uVs 16:18:40 Integrated Voltage Measurement... 16:18:49 Voltage Offset Correction: v_off_ini = 1100.670 uV, v_off_fin = 1096.922 uV 16:18:49 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 227.910 uVs 16:18:49 Integrated Voltage Measurement... 16:18:58 Voltage Offset Correction: v_off_ini = 1083.349 uV, v_off_fin = 1068.166 uV 16:18:58 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -246.758 uVs 16:18:58 Integrated Voltage Measurement... 16:19:07 Voltage Offset Correction: v_off_ini = 1089.990 uV, v_off_fin = 1111.471 uV 16:19:07 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 253.848 uVs 16:19:07 Integrated Voltage Measurement... 16:19:16 Voltage Offset Correction: v_off_ini = 1084.283 uV, v_off_fin = 1082.139 uV 16:19:16 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -253.201 uVs 16:19:16 Integrated Voltage Measurement... 16:19:25 Voltage Offset Correction: v_off_ini = 1091.012 uV, v_off_fin = 1102.182 uV 16:19:25 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 244.273 uVs 16:19:29 I1Y Average and RMS of Measurements VT = -247.667 +- 2.698 uVs I1Y = -49.533 +- 0.540 uTm 16:19:34 Wire Position: X0 = -2.0000 mm, Y0 = 0.0000 mm 16:19:37 Integrated Voltage Measurement... 16:19:46 Voltage Offset Correction: v_off_ini = 1093.729 uV, v_off_fin = 1091.111 uV 16:19:46 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -93.030 uVs 16:19:46 Integrated Voltage Measurement... 16:19:55 Voltage Offset Correction: v_off_ini = 1087.173 uV, v_off_fin = 1099.419 uV 16:19:55 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 103.342 uVs 16:19:55 Integrated Voltage Measurement... 16:20:04 Voltage Offset Correction: v_off_ini = 1113.293 uV, v_off_fin = 1099.785 uV 16:20:04 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -117.085 uVs 16:20:04 Integrated Voltage Measurement... 16:20:13 Voltage Offset Correction: v_off_ini = 1109.960 uV, v_off_fin = 1128.084 uV 16:20:13 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 98.306 uVs 16:20:13 Integrated Voltage Measurement... 16:20:22 Voltage Offset Correction: v_off_ini = 1104.498 uV, v_off_fin = 1083.579 uV 16:20:22 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -94.921 uVs 16:20:22 Integrated Voltage Measurement... 16:20:31 Voltage Offset Correction: v_off_ini = 1109.665 uV, v_off_fin = 1110.272 uV 16:20:31 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 70.483 uVs 16:20:33 I2X Average and RMS of Measurements VT = -96.194 +- 10.300 uVs I2X = 8.850 +- 0.948 uTm^2 16:20:38 Wire Position: X0 = -2.0000 mm, Y0 = 0.0000 mm 16:20:41 Integrated Voltage Measurement... 16:20:50 Voltage Offset Correction: v_off_ini = 1116.674 uV, v_off_fin = 1102.376 uV 16:20:50 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -269.250 uVs 16:20:50 Integrated Voltage Measurement... 16:20:59 Voltage Offset Correction: v_off_ini = 1105.545 uV, v_off_fin = 1104.560 uV 16:20:59 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 282.458 uVs 16:20:59 Integrated Voltage Measurement... 16:21:08 Voltage Offset Correction: v_off_ini = 1111.065 uV, v_off_fin = 1105.059 uV 16:21:08 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -285.631 uVs 16:21:08 Integrated Voltage Measurement... 16:21:17 Voltage Offset Correction: v_off_ini = 1093.625 uV, v_off_fin = 1124.595 uV 16:21:17 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 279.050 uVs 16:21:17 Integrated Voltage Measurement... 16:21:26 Voltage Offset Correction: v_off_ini = 1103.885 uV, v_off_fin = 1107.750 uV 16:21:26 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -267.368 uVs 16:21:26 Integrated Voltage Measurement... 16:21:35 Voltage Offset Correction: v_off_ini = 1123.585 uV, v_off_fin = 1119.720 uV 16:21:35 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 290.568 uVs 16:21:38 I2Y Average and RMS of Measurements VT = -279.054 +- 2.649 uVs I2Y = -25.673 +- 0.244 uTm^2 16:21:42 Wire Position: X0 = -1.0000 mm, Y0 = 0.0000 mm 16:21:46 Integrated Voltage Measurement... 16:21:55 Voltage Offset Correction: v_off_ini = 1125.194 uV, v_off_fin = 1116.208 uV 16:21:55 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -99.912 uVs 16:21:56 Integrated Voltage Measurement... 16:22:04 Voltage Offset Correction: v_off_ini = 1099.912 uV, v_off_fin = 1102.365 uV 16:22:04 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 86.916 uVs 16:22:05 Integrated Voltage Measurement... 16:22:13 Voltage Offset Correction: v_off_ini = 1109.901 uV, v_off_fin = 1104.648 uV 16:22:14 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -82.101 uVs 16:22:14 Integrated Voltage Measurement... 16:22:22 Voltage Offset Correction: v_off_ini = 1096.795 uV, v_off_fin = 1109.993 uV 16:22:23 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 102.002 uVs 16:22:23 Integrated Voltage Measurement... 16:22:32 Voltage Offset Correction: v_off_ini = 1123.619 uV, v_off_fin = 1120.664 uV 16:22:32 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -70.293 uVs 16:22:32 Integrated Voltage Measurement... 16:22:41 Voltage Offset Correction: v_off_ini = 1103.850 uV, v_off_fin = 1108.276 uV 16:22:41 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 82.406 uVs 16:22:45 I1X Average and RMS of Measurements VT = -87.272 +- 7.743 uVs I1X = 17.454 +- 1.549 uTm 16:22:47 Wire Position: X0 = -1.0000 mm, Y0 = 0.0000 mm 16:22:52 Integrated Voltage Measurement... 16:23:01 Voltage Offset Correction: v_off_ini = 1098.504 uV, v_off_fin = 1092.245 uV 16:23:01 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -234.141 uVs 16:23:01 Integrated Voltage Measurement... 16:23:10 Voltage Offset Correction: v_off_ini = 1120.824 uV, v_off_fin = 1125.309 uV 16:23:10 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 244.102 uVs 16:23:10 Integrated Voltage Measurement... 16:23:19 Voltage Offset Correction: v_off_ini = 1124.601 uV, v_off_fin = 1104.463 uV 16:23:19 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -246.320 uVs 16:23:19 Integrated Voltage Measurement... 16:23:28 Voltage Offset Correction: v_off_ini = 1114.758 uV, v_off_fin = 1110.294 uV 16:23:28 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 250.584 uVs 16:23:28 Integrated Voltage Measurement... 16:23:37 Voltage Offset Correction: v_off_ini = 1102.115 uV, v_off_fin = 1102.963 uV 16:23:37 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -246.445 uVs 16:23:37 Integrated Voltage Measurement... 16:23:46 Voltage Offset Correction: v_off_ini = 1100.158 uV, v_off_fin = 1123.652 uV 16:23:46 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 231.964 uVs 16:23:50 I1Y Average and RMS of Measurements VT = -242.259 +- 4.379 uVs I1Y = -48.452 +- 0.876 uTm 16:23:55 Wire Position: X0 = -1.0000 mm, Y0 = 0.0000 mm 16:23:58 Integrated Voltage Measurement... 16:24:07 Voltage Offset Correction: v_off_ini = 1086.832 uV, v_off_fin = 1087.949 uV 16:24:07 Integrated Voltage Measurement dXYus = -1.0001 mm, dXYds = 0.0000 mm dXY = -1.0001 mm VT = -66.504 uVs 16:24:07 Integrated Voltage Measurement... 16:24:16 Voltage Offset Correction: v_off_ini = 1110.167 uV, v_off_fin = 1109.083 uV 16:24:16 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 77.050 uVs 16:24:16 Integrated Voltage Measurement... 16:24:25 Voltage Offset Correction: v_off_ini = 1113.721 uV, v_off_fin = 1105.135 uV 16:24:25 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -81.678 uVs 16:24:25 Integrated Voltage Measurement... 16:24:34 Voltage Offset Correction: v_off_ini = 1114.318 uV, v_off_fin = 1104.454 uV 16:24:34 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 47.499 uVs 16:24:34 Integrated Voltage Measurement... 16:24:43 Voltage Offset Correction: v_off_ini = 1101.202 uV, v_off_fin = 1098.607 uV 16:24:43 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -73.288 uVs 16:24:43 Integrated Voltage Measurement... 16:24:52 Voltage Offset Correction: v_off_ini = 1127.237 uV, v_off_fin = 1129.789 uV 16:24:52 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 78.756 uVs 16:24:54 I2X Average and RMS of Measurements VT = -70.796 +- 4.719 uVs I2X = 6.513 +- 0.434 uTm^2 16:24:59 Wire Position: X0 = -1.0000 mm, Y0 = 0.0000 mm 16:25:02 Integrated Voltage Measurement... 16:25:11 Voltage Offset Correction: v_off_ini = 1123.211 uV, v_off_fin = 1107.569 uV 16:25:11 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -253.303 uVs 16:25:11 Integrated Voltage Measurement... 16:25:20 Voltage Offset Correction: v_off_ini = 1110.636 uV, v_off_fin = 1120.182 uV 16:25:20 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 261.666 uVs 16:25:20 Integrated Voltage Measurement... 16:25:29 Voltage Offset Correction: v_off_ini = 1107.342 uV, v_off_fin = 1093.418 uV 16:25:29 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -246.036 uVs 16:25:29 Integrated Voltage Measurement... 16:25:38 Voltage Offset Correction: v_off_ini = 1095.799 uV, v_off_fin = 1094.399 uV 16:25:38 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 265.012 uVs 16:25:38 Integrated Voltage Measurement... 16:25:47 Voltage Offset Correction: v_off_ini = 1085.850 uV, v_off_fin = 1096.482 uV 16:25:47 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -269.552 uVs 16:25:47 Integrated Voltage Measurement... 16:25:56 Voltage Offset Correction: v_off_ini = 1093.234 uV, v_off_fin = 1104.798 uV 16:25:56 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 256.985 uVs 16:25:58 I2Y Average and RMS of Measurements VT = -258.759 +- 3.288 uVs I2Y = -23.806 +- 0.302 uTm^2 16:26:03 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:26:07 Integrated Voltage Measurement... 16:26:16 Voltage Offset Correction: v_off_ini = 1113.000 uV, v_off_fin = 1112.200 uV 16:26:16 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -63.024 uVs 16:26:16 Integrated Voltage Measurement... 16:26:25 Voltage Offset Correction: v_off_ini = 1110.049 uV, v_off_fin = 1100.152 uV 16:26:25 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 72.358 uVs 16:26:25 Integrated Voltage Measurement... 16:26:34 Voltage Offset Correction: v_off_ini = 1091.614 uV, v_off_fin = 1107.032 uV 16:26:34 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -57.822 uVs 16:26:35 Integrated Voltage Measurement... 16:26:43 Voltage Offset Correction: v_off_ini = 1116.021 uV, v_off_fin = 1107.932 uV 16:26:44 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 77.007 uVs 16:26:44 Integrated Voltage Measurement... 16:26:53 Voltage Offset Correction: v_off_ini = 1085.022 uV, v_off_fin = 1091.158 uV 16:26:53 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -45.161 uVs 16:26:53 Integrated Voltage Measurement... 16:27:02 Voltage Offset Correction: v_off_ini = 1075.223 uV, v_off_fin = 1094.389 uV 16:27:02 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 91.655 uVs 16:27:06 I1X Average and RMS of Measurements VT = -67.838 +- 0.419 uVs I1X = 13.568 +- 0.084 uTm 16:27:08 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:27:13 Integrated Voltage Measurement... 16:27:22 Voltage Offset Correction: v_off_ini = 1094.306 uV, v_off_fin = 1070.600 uV 16:27:22 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = -256.880 uVs 16:27:22 Integrated Voltage Measurement... 16:27:31 Voltage Offset Correction: v_off_ini = 1101.701 uV, v_off_fin = 1089.038 uV 16:27:31 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 246.874 uVs 16:27:31 Integrated Voltage Measurement... 16:27:40 Voltage Offset Correction: v_off_ini = 1053.549 uV, v_off_fin = 1045.716 uV 16:27:40 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -250.133 uVs 16:27:40 Integrated Voltage Measurement... 16:27:49 Voltage Offset Correction: v_off_ini = 1061.851 uV, v_off_fin = 1068.970 uV 16:27:49 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 245.559 uVs 16:27:49 Integrated Voltage Measurement... 16:27:58 Voltage Offset Correction: v_off_ini = 1058.691 uV, v_off_fin = 1056.425 uV 16:27:58 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -248.598 uVs 16:27:58 Integrated Voltage Measurement... 16:28:07 Voltage Offset Correction: v_off_ini = 1049.820 uV, v_off_fin = 1076.933 uV 16:28:07 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 231.846 uVs 16:28:11 I1Y Average and RMS of Measurements VT = -246.648 +- 4.833 uVs I1Y = -49.331 +- 0.968 uTm 16:28:16 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:28:19 Integrated Voltage Measurement... 16:28:28 Voltage Offset Correction: v_off_ini = 1053.386 uV, v_off_fin = 1026.668 uV 16:28:28 Integrated Voltage Measurement dXYus = -1.0001 mm, dXYds = 0.0000 mm dXY = -1.0001 mm VT = -76.229 uVs 16:28:28 Integrated Voltage Measurement... 16:28:37 Voltage Offset Correction: v_off_ini = 1065.053 uV, v_off_fin = 1062.637 uV 16:28:37 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 73.276 uVs 16:28:37 Integrated Voltage Measurement... 16:28:46 Voltage Offset Correction: v_off_ini = 1051.549 uV, v_off_fin = 1052.502 uV 16:28:46 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -79.415 uVs 16:28:46 Integrated Voltage Measurement... 16:28:55 Voltage Offset Correction: v_off_ini = 1044.033 uV, v_off_fin = 1045.028 uV 16:28:55 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 116.618 uVs 16:28:55 Integrated Voltage Measurement... 16:29:04 Voltage Offset Correction: v_off_ini = 1031.681 uV, v_off_fin = 1044.011 uV 16:29:04 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -87.683 uVs 16:29:04 Integrated Voltage Measurement... 16:29:13 Voltage Offset Correction: v_off_ini = 1033.804 uV, v_off_fin = 1032.851 uV 16:29:13 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 90.755 uVs 16:29:15 I2X Average and RMS of Measurements VT = -87.329 +- 9.591 uVs I2X = 8.034 +- 0.883 uTm^2 16:29:20 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:29:23 Integrated Voltage Measurement... 16:29:32 Voltage Offset Correction: v_off_ini = 1045.041 uV, v_off_fin = 1036.035 uV 16:29:32 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -265.609 uVs 16:29:32 Integrated Voltage Measurement... 16:29:41 Voltage Offset Correction: v_off_ini = 1020.715 uV, v_off_fin = 1024.529 uV 16:29:41 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 280.934 uVs 16:29:41 Integrated Voltage Measurement... 16:29:50 Voltage Offset Correction: v_off_ini = 1018.742 uV, v_off_fin = 1024.089 uV 16:29:50 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -245.893 uVs 16:29:50 Integrated Voltage Measurement... 16:29:59 Voltage Offset Correction: v_off_ini = 1030.869 uV, v_off_fin = 1026.532 uV 16:29:59 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 249.238 uVs 16:29:59 Integrated Voltage Measurement... 16:30:08 Voltage Offset Correction: v_off_ini = 1042.640 uV, v_off_fin = 1044.607 uV 16:30:08 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -251.229 uVs 16:30:08 Integrated Voltage Measurement... 16:30:17 Voltage Offset Correction: v_off_ini = 1048.069 uV, v_off_fin = 1028.262 uV 16:30:17 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 287.389 uVs 16:30:20 I2Y Average and RMS of Measurements VT = -263.382 +- 11.300 uVs I2Y = -24.231 +- 1.040 uTm^2 16:30:24 Wire Position: X0 = 1.0000 mm, Y0 = 0.0000 mm 16:30:28 Integrated Voltage Measurement... 16:30:37 Voltage Offset Correction: v_off_ini = 1038.250 uV, v_off_fin = 1024.978 uV 16:30:37 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -59.514 uVs 16:30:37 Integrated Voltage Measurement... 16:30:46 Voltage Offset Correction: v_off_ini = 1016.898 uV, v_off_fin = 1007.600 uV 16:30:46 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 77.378 uVs 16:30:47 Integrated Voltage Measurement... 16:30:55 Voltage Offset Correction: v_off_ini = 1013.816 uV, v_off_fin = 1027.589 uV 16:30:55 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -63.608 uVs 16:30:56 Integrated Voltage Measurement... 16:31:04 Voltage Offset Correction: v_off_ini = 1040.214 uV, v_off_fin = 1031.543 uV 16:31:05 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 82.433 uVs 16:31:05 Integrated Voltage Measurement... 16:31:14 Voltage Offset Correction: v_off_ini = 1020.724 uV, v_off_fin = 1040.479 uV 16:31:14 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -75.061 uVs 16:31:14 Integrated Voltage Measurement... 16:31:23 Voltage Offset Correction: v_off_ini = 1029.855 uV, v_off_fin = 1030.135 uV 16:31:23 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 52.814 uVs 16:31:27 I1X Average and RMS of Measurements VT = -68.468 +- 3.708 uVs I1X = 13.694 +- 0.742 uTm 16:31:29 Wire Position: X0 = 1.0000 mm, Y0 = 0.0000 mm 16:31:34 Integrated Voltage Measurement... 16:31:42 Voltage Offset Correction: v_off_ini = 1041.881 uV, v_off_fin = 1035.960 uV 16:31:43 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -249.216 uVs 16:31:43 Integrated Voltage Measurement... 16:31:51 Voltage Offset Correction: v_off_ini = 1043.558 uV, v_off_fin = 1048.328 uV 16:31:52 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 245.529 uVs 16:31:52 Integrated Voltage Measurement... 16:32:00 Voltage Offset Correction: v_off_ini = 1042.216 uV, v_off_fin = 1028.079 uV 16:32:01 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -244.892 uVs 16:32:01 Integrated Voltage Measurement... 16:32:09 Voltage Offset Correction: v_off_ini = 1042.048 uV, v_off_fin = 1037.146 uV 16:32:10 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 228.568 uVs 16:32:10 Integrated Voltage Measurement... 16:32:19 Voltage Offset Correction: v_off_ini = 1025.966 uV, v_off_fin = 1005.834 uV 16:32:19 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -228.017 uVs 16:32:19 Integrated Voltage Measurement... 16:32:28 Voltage Offset Correction: v_off_ini = 1026.119 uV, v_off_fin = 1036.144 uV 16:32:28 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 246.597 uVs 16:32:32 I1Y Average and RMS of Measurements VT = -240.470 +- 4.887 uVs I1Y = -48.094 +- 0.977 uTm 16:32:37 Wire Position: X0 = 1.0000 mm, Y0 = 0.0000 mm 16:32:40 Integrated Voltage Measurement... 16:32:49 Voltage Offset Correction: v_off_ini = 1021.863 uV, v_off_fin = 1008.894 uV 16:32:49 Integrated Voltage Measurement dXYus = -1.0001 mm, dXYds = 0.0000 mm dXY = -1.0001 mm VT = -78.314 uVs 16:32:49 Integrated Voltage Measurement... 16:32:58 Voltage Offset Correction: v_off_ini = 1007.758 uV, v_off_fin = 1013.049 uV 16:32:58 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 71.790 uVs 16:32:58 Integrated Voltage Measurement... 16:33:07 Voltage Offset Correction: v_off_ini = 1022.245 uV, v_off_fin = 1018.017 uV 16:33:07 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -75.468 uVs 16:33:07 Integrated Voltage Measurement... 16:33:16 Voltage Offset Correction: v_off_ini = 1020.237 uV, v_off_fin = 1017.194 uV 16:33:16 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 68.720 uVs 16:33:16 Integrated Voltage Measurement... 16:33:25 Voltage Offset Correction: v_off_ini = 1005.867 uV, v_off_fin = 1011.200 uV 16:33:25 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -73.414 uVs 16:33:25 Integrated Voltage Measurement... 16:33:34 Voltage Offset Correction: v_off_ini = 998.211 uV, v_off_fin = 1011.562 uV 16:33:34 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 80.843 uVs 16:33:36 I2X Average and RMS of Measurements VT = -74.758 +- 2.066 uVs I2X = 6.878 +- 0.190 uTm^2 16:33:41 Wire Position: X0 = 1.0000 mm, Y0 = 0.0000 mm 16:33:44 Integrated Voltage Measurement... 16:33:53 Voltage Offset Correction: v_off_ini = 1012.764 uV, v_off_fin = 1023.288 uV 16:33:53 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -266.459 uVs 16:33:53 Integrated Voltage Measurement... 16:34:02 Voltage Offset Correction: v_off_ini = 1017.679 uV, v_off_fin = 1024.041 uV 16:34:02 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 263.851 uVs 16:34:02 Integrated Voltage Measurement... 16:34:11 Voltage Offset Correction: v_off_ini = 1010.304 uV, v_off_fin = 1036.759 uV 16:34:11 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -286.209 uVs 16:34:11 Integrated Voltage Measurement... 16:34:20 Voltage Offset Correction: v_off_ini = 1053.204 uV, v_off_fin = 1028.656 uV 16:34:20 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 234.018 uVs 16:34:20 Integrated Voltage Measurement... 16:34:29 Voltage Offset Correction: v_off_ini = 1000.884 uV, v_off_fin = 999.757 uV 16:34:29 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -281.514 uVs 16:34:29 Integrated Voltage Measurement... 16:34:38 Voltage Offset Correction: v_off_ini = 1003.099 uV, v_off_fin = 1016.633 uV 16:34:38 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 276.114 uVs 16:34:40 I2Y Average and RMS of Measurements VT = -268.028 +- 7.900 uVs I2Y = -24.659 +- 0.727 uTm^2 16:34:45 Wire Position: X0 = 2.0000 mm, Y0 = 0.0000 mm 16:34:49 Integrated Voltage Measurement... 16:34:58 Voltage Offset Correction: v_off_ini = 984.269 uV, v_off_fin = 992.298 uV 16:34:58 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -60.361 uVs 16:34:58 Integrated Voltage Measurement... 16:35:07 Voltage Offset Correction: v_off_ini = 989.180 uV, v_off_fin = 989.181 uV 16:35:07 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 89.023 uVs 16:35:07 Integrated Voltage Measurement... 16:35:16 Voltage Offset Correction: v_off_ini = 988.844 uV, v_off_fin = 1007.409 uV 16:35:16 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -76.701 uVs 16:35:16 Integrated Voltage Measurement... 16:35:25 Voltage Offset Correction: v_off_ini = 990.750 uV, v_off_fin = 993.070 uV 16:35:25 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 52.121 uVs 16:35:25 Integrated Voltage Measurement... 16:35:34 Voltage Offset Correction: v_off_ini = 987.712 uV, v_off_fin = 994.577 uV 16:35:34 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -68.470 uVs 16:35:34 Integrated Voltage Measurement... 16:35:43 Voltage Offset Correction: v_off_ini = 998.538 uV, v_off_fin = 1002.959 uV 16:35:43 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 69.222 uVs 16:35:48 I1X Average and RMS of Measurements VT = -69.316 +- 4.210 uVs I1X = 13.863 +- 0.842 uTm 16:35:50 Wire Position: X0 = 2.0000 mm, Y0 = 0.0000 mm 16:35:54 Integrated Voltage Measurement... 16:36:03 Voltage Offset Correction: v_off_ini = 1007.040 uV, v_off_fin = 994.582 uV 16:36:03 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -229.736 uVs 16:36:03 Integrated Voltage Measurement... 16:36:12 Voltage Offset Correction: v_off_ini = 1012.941 uV, v_off_fin = 1030.880 uV 16:36:12 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 241.022 uVs 16:36:12 Integrated Voltage Measurement... 16:36:21 Voltage Offset Correction: v_off_ini = 1005.959 uV, v_off_fin = 986.101 uV 16:36:21 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -249.655 uVs 16:36:21 Integrated Voltage Measurement... 16:36:30 Voltage Offset Correction: v_off_ini = 1014.401 uV, v_off_fin = 1018.361 uV 16:36:30 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 239.853 uVs 16:36:30 Integrated Voltage Measurement... 16:36:39 Voltage Offset Correction: v_off_ini = 1005.405 uV, v_off_fin = 989.867 uV 16:36:39 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -249.576 uVs 16:36:39 Integrated Voltage Measurement... 16:36:48 Voltage Offset Correction: v_off_ini = 991.783 uV, v_off_fin = 989.674 uV 16:36:48 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 239.116 uVs 16:36:53 I1Y Average and RMS of Measurements VT = -241.493 +- 4.326 uVs I1Y = -48.299 +- 0.865 uTm 16:36:58 Wire Position: X0 = 2.0000 mm, Y0 = 0.0000 mm 16:37:00 Integrated Voltage Measurement... 16:37:09 Voltage Offset Correction: v_off_ini = 1006.729 uV, v_off_fin = 1012.764 uV 16:37:09 Integrated Voltage Measurement dXYus = -1.0001 mm, dXYds = 0.0000 mm dXY = -1.0001 mm VT = -78.168 uVs 16:37:09 Integrated Voltage Measurement... 16:37:18 Voltage Offset Correction: v_off_ini = 1002.016 uV, v_off_fin = 1009.494 uV 16:37:18 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 70.168 uVs 16:37:18 Integrated Voltage Measurement... 16:37:27 Voltage Offset Correction: v_off_ini = 996.988 uV, v_off_fin = 1004.390 uV 16:37:27 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -53.021 uVs 16:37:28 Integrated Voltage Measurement... 16:37:36 Voltage Offset Correction: v_off_ini = 1020.898 uV, v_off_fin = 1030.822 uV 16:37:37 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 68.688 uVs 16:37:37 Integrated Voltage Measurement... 16:37:46 Voltage Offset Correction: v_off_ini = 1015.364 uV, v_off_fin = 1016.717 uV 16:37:46 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -83.718 uVs 16:37:46 Integrated Voltage Measurement... 16:37:55 Voltage Offset Correction: v_off_ini = 1027.123 uV, v_off_fin = 1034.363 uV 16:37:55 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 63.560 uVs 16:37:57 I2X Average and RMS of Measurements VT = -69.554 +- 6.155 uVs I2X = 6.399 +- 0.566 uTm^2 16:38:02 Wire Position: X0 = 2.0000 mm, Y0 = 0.0000 mm 16:38:05 Integrated Voltage Measurement... 16:38:14 Voltage Offset Correction: v_off_ini = 1012.906 uV, v_off_fin = 1035.294 uV 16:38:14 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -275.198 uVs 16:38:14 Integrated Voltage Measurement... 16:38:23 Voltage Offset Correction: v_off_ini = 1037.517 uV, v_off_fin = 1008.996 uV 16:38:23 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 276.895 uVs 16:38:23 Integrated Voltage Measurement... 16:38:32 Voltage Offset Correction: v_off_ini = 1026.216 uV, v_off_fin = 1015.188 uV 16:38:32 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -260.734 uVs 16:38:32 Integrated Voltage Measurement... 16:38:41 Voltage Offset Correction: v_off_ini = 1022.463 uV, v_off_fin = 1016.305 uV 16:38:41 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 257.983 uVs 16:38:41 Integrated Voltage Measurement... 16:38:50 Voltage Offset Correction: v_off_ini = 1020.282 uV, v_off_fin = 1013.988 uV 16:38:50 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -261.546 uVs 16:38:50 Integrated Voltage Measurement... 16:38:59 Voltage Offset Correction: v_off_ini = 1005.315 uV, v_off_fin = 1017.872 uV 16:38:59 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 271.490 uVs 16:39:01 I2Y Average and RMS of Measurements VT = -267.308 +- 6.836 uVs I2Y = -24.592 +- 0.629 uTm^2 16:39:09 Wire Position: X0 = 0.0000 mm, Y0 = -1.0000 mm 16:39:13 Integrated Voltage Measurement... 16:39:22 Voltage Offset Correction: v_off_ini = 1021.237 uV, v_off_fin = 1007.315 uV 16:39:22 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -84.606 uVs 16:39:22 Integrated Voltage Measurement... 16:39:31 Voltage Offset Correction: v_off_ini = 1031.383 uV, v_off_fin = 1014.351 uV 16:39:31 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 83.324 uVs 16:39:31 Integrated Voltage Measurement... 16:39:40 Voltage Offset Correction: v_off_ini = 1026.507 uV, v_off_fin = 1005.058 uV 16:39:40 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -83.408 uVs 16:39:40 Integrated Voltage Measurement... 16:39:49 Voltage Offset Correction: v_off_ini = 1021.116 uV, v_off_fin = 1032.944 uV 16:39:49 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 102.178 uVs 16:39:50 Integrated Voltage Measurement... 16:39:58 Voltage Offset Correction: v_off_ini = 1033.573 uV, v_off_fin = 1038.336 uV 16:39:58 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -82.578 uVs 16:39:59 Integrated Voltage Measurement... 16:40:07 Voltage Offset Correction: v_off_ini = 1039.514 uV, v_off_fin = 1041.787 uV 16:40:08 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 70.102 uVs 16:40:12 I1X Average and RMS of Measurements VT = -84.366 +- 6.723 uVs I1X = 16.873 +- 1.345 uTm 16:40:14 Wire Position: X0 = 0.0000 mm, Y0 = -1.0000 mm 16:40:18 Integrated Voltage Measurement... 16:40:27 Voltage Offset Correction: v_off_ini = 1053.359 uV, v_off_fin = 1032.436 uV 16:40:27 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -235.019 uVs 16:40:28 Integrated Voltage Measurement... 16:40:36 Voltage Offset Correction: v_off_ini = 1032.120 uV, v_off_fin = 1060.534 uV 16:40:36 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 248.715 uVs 16:40:37 Integrated Voltage Measurement... 16:40:45 Voltage Offset Correction: v_off_ini = 1044.339 uV, v_off_fin = 1008.571 uV 16:40:45 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -247.698 uVs 16:40:46 Integrated Voltage Measurement... 16:40:54 Voltage Offset Correction: v_off_ini = 1035.676 uV, v_off_fin = 1038.622 uV 16:40:54 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 248.146 uVs 16:40:55 Integrated Voltage Measurement... 16:41:03 Voltage Offset Correction: v_off_ini = 1022.617 uV, v_off_fin = 1021.981 uV 16:41:03 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -230.378 uVs 16:41:04 Integrated Voltage Measurement... 16:41:12 Voltage Offset Correction: v_off_ini = 1031.964 uV, v_off_fin = 1029.400 uV 16:41:13 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 252.761 uVs 16:41:17 I1Y Average and RMS of Measurements VT = -243.786 +- 2.927 uVs I1Y = -48.757 +- 0.585 uTm 16:41:19 Wire Position: X0 = 0.0000 mm, Y0 = -1.0000 mm 16:41:22 Integrated Voltage Measurement... 16:41:31 Voltage Offset Correction: v_off_ini = 1020.534 uV, v_off_fin = 1022.959 uV 16:41:31 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -110.130 uVs 16:41:31 Integrated Voltage Measurement... 16:41:40 Voltage Offset Correction: v_off_ini = 1024.390 uV, v_off_fin = 1039.357 uV 16:41:40 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 118.042 uVs 16:41:40 Integrated Voltage Measurement... 16:41:49 Voltage Offset Correction: v_off_ini = 1021.377 uV, v_off_fin = 1006.301 uV 16:41:49 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -102.788 uVs 16:41:49 Integrated Voltage Measurement... 16:41:58 Voltage Offset Correction: v_off_ini = 1004.987 uV, v_off_fin = 1024.048 uV 16:41:58 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 106.270 uVs 16:41:58 Integrated Voltage Measurement... 16:42:07 Voltage Offset Correction: v_off_ini = 1013.190 uV, v_off_fin = 990.334 uV 16:42:07 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -116.289 uVs 16:42:07 Integrated Voltage Measurement... 16:42:16 Voltage Offset Correction: v_off_ini = 987.919 uV, v_off_fin = 1000.392 uV 16:42:16 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 106.145 uVs 16:42:18 I2X Average and RMS of Measurements VT = -109.944 +- 4.004 uVs I2X = 10.115 +- 0.369 uTm^2 16:42:21 Wire Position: X0 = 0.0000 mm, Y0 = -1.0000 mm 16:42:24 Integrated Voltage Measurement... 16:42:33 Voltage Offset Correction: v_off_ini = 1031.662 uV, v_off_fin = 1030.917 uV 16:42:33 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -271.850 uVs 16:42:33 Integrated Voltage Measurement... 16:42:42 Voltage Offset Correction: v_off_ini = 1024.336 uV, v_off_fin = 1014.152 uV 16:42:42 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 279.764 uVs 16:42:42 Integrated Voltage Measurement... 16:42:51 Voltage Offset Correction: v_off_ini = 994.035 uV, v_off_fin = 1016.865 uV 16:42:51 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -279.060 uVs 16:42:51 Integrated Voltage Measurement... 16:43:00 Voltage Offset Correction: v_off_ini = 998.682 uV, v_off_fin = 1010.293 uV 16:43:00 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 275.294 uVs 16:43:00 Integrated Voltage Measurement... 16:43:09 Voltage Offset Correction: v_off_ini = 1006.198 uV, v_off_fin = 999.068 uV 16:43:09 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -282.036 uVs 16:43:09 Integrated Voltage Measurement... 16:43:18 Voltage Offset Correction: v_off_ini = 1015.038 uV, v_off_fin = 1013.127 uV 16:43:18 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 267.125 uVs 16:43:20 I2Y Average and RMS of Measurements VT = -275.855 +- 1.061 uVs I2Y = -25.379 +- 0.098 uTm^2 16:43:24 Wire Position: X0 = 0.0000 mm, Y0 = -0.5000 mm 16:43:29 Integrated Voltage Measurement... 16:43:38 Voltage Offset Correction: v_off_ini = 1029.610 uV, v_off_fin = 1015.449 uV 16:43:38 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -76.768 uVs 16:43:38 Integrated Voltage Measurement... 16:43:47 Voltage Offset Correction: v_off_ini = 1027.291 uV, v_off_fin = 1022.384 uV 16:43:47 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 61.998 uVs 16:43:47 Integrated Voltage Measurement... 16:43:56 Voltage Offset Correction: v_off_ini = 1006.411 uV, v_off_fin = 1013.723 uV 16:43:56 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -95.267 uVs 16:43:56 Integrated Voltage Measurement... 16:44:05 Voltage Offset Correction: v_off_ini = 998.553 uV, v_off_fin = 1018.418 uV 16:44:05 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 81.692 uVs 16:44:05 Integrated Voltage Measurement... 16:44:14 Voltage Offset Correction: v_off_ini = 1034.127 uV, v_off_fin = 1020.100 uV 16:44:14 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -63.690 uVs 16:44:14 Integrated Voltage Measurement... 16:44:23 Voltage Offset Correction: v_off_ini = 1010.594 uV, v_off_fin = 1022.594 uV 16:44:23 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 71.645 uVs 16:44:27 I1X Average and RMS of Measurements VT = -75.177 +- 9.433 uVs I1X = 15.036 +- 1.886 uTm 16:44:29 Wire Position: X0 = 0.0000 mm, Y0 = -0.5000 mm 16:44:34 Integrated Voltage Measurement... 16:44:43 Voltage Offset Correction: v_off_ini = 1045.173 uV, v_off_fin = 1022.406 uV 16:44:43 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -255.993 uVs 16:44:43 Integrated Voltage Measurement... 16:44:52 Voltage Offset Correction: v_off_ini = 1023.502 uV, v_off_fin = 1042.225 uV 16:44:52 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 235.322 uVs 16:44:52 Integrated Voltage Measurement... 16:45:01 Voltage Offset Correction: v_off_ini = 1007.731 uV, v_off_fin = 1009.607 uV 16:45:01 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -236.777 uVs 16:45:01 Integrated Voltage Measurement... 16:45:10 Voltage Offset Correction: v_off_ini = 1016.004 uV, v_off_fin = 1034.964 uV 16:45:10 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 241.551 uVs 16:45:10 Integrated Voltage Measurement... 16:45:19 Voltage Offset Correction: v_off_ini = 1031.226 uV, v_off_fin = 1019.466 uV 16:45:19 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -247.747 uVs 16:45:19 Integrated Voltage Measurement... 16:45:28 Voltage Offset Correction: v_off_ini = 1043.868 uV, v_off_fin = 1043.951 uV 16:45:28 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 256.501 uVs 16:45:32 I1Y Average and RMS of Measurements VT = -245.648 +- 5.291 uVs I1Y = -49.130 +- 1.058 uTm 16:45:34 Wire Position: X0 = 0.0000 mm, Y0 = -0.5000 mm 16:45:37 Integrated Voltage Measurement... 16:45:46 Voltage Offset Correction: v_off_ini = 1020.326 uV, v_off_fin = 1024.377 uV 16:45:46 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -82.074 uVs 16:45:46 Integrated Voltage Measurement... 16:45:55 Voltage Offset Correction: v_off_ini = 1043.807 uV, v_off_fin = 1044.355 uV 16:45:55 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 74.741 uVs 16:45:55 Integrated Voltage Measurement... 16:46:04 Voltage Offset Correction: v_off_ini = 1017.743 uV, v_off_fin = 1011.006 uV 16:46:04 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -72.416 uVs 16:46:04 Integrated Voltage Measurement... 16:46:13 Voltage Offset Correction: v_off_ini = 1013.583 uV, v_off_fin = 1024.349 uV 16:46:13 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 109.133 uVs 16:46:13 Integrated Voltage Measurement... 16:46:22 Voltage Offset Correction: v_off_ini = 1001.791 uV, v_off_fin = 1009.831 uV 16:46:22 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -84.106 uVs 16:46:22 Integrated Voltage Measurement... 16:46:31 Voltage Offset Correction: v_off_ini = 1033.498 uV, v_off_fin = 1011.502 uV 16:46:31 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 73.859 uVs 16:46:34 I2X Average and RMS of Measurements VT = -82.722 +- 5.699 uVs I2X = 7.610 +- 0.524 uTm^2 16:46:36 Wire Position: X0 = 0.0000 mm, Y0 = -0.5000 mm 16:46:39 Integrated Voltage Measurement... 16:46:48 Voltage Offset Correction: v_off_ini = 971.455 uV, v_off_fin = 995.751 uV 16:46:48 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -228.927 uVs 16:46:48 Integrated Voltage Measurement... 16:46:57 Voltage Offset Correction: v_off_ini = 999.551 uV, v_off_fin = 1004.545 uV 16:46:57 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 239.359 uVs 16:46:57 Integrated Voltage Measurement... 16:47:06 Voltage Offset Correction: v_off_ini = 992.192 uV, v_off_fin = 993.945 uV 16:47:06 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -249.256 uVs 16:47:06 Integrated Voltage Measurement... 16:47:15 Voltage Offset Correction: v_off_ini = 995.491 uV, v_off_fin = 1024.172 uV 16:47:15 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 278.712 uVs 16:47:15 Integrated Voltage Measurement... 16:47:24 Voltage Offset Correction: v_off_ini = 1008.298 uV, v_off_fin = 1001.936 uV 16:47:24 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = -273.627 uVs 16:47:24 Integrated Voltage Measurement... 16:47:33 Voltage Offset Correction: v_off_ini = 1000.432 uV, v_off_fin = 1005.795 uV 16:47:33 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 268.281 uVs 16:47:35 I2Y Average and RMS of Measurements VT = -256.360 +- 15.966 uVs I2Y = -23.586 +- 1.469 uTm^2 16:47:39 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:47:44 Integrated Voltage Measurement... 16:47:53 Voltage Offset Correction: v_off_ini = 1003.169 uV, v_off_fin = 987.874 uV 16:47:53 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -72.558 uVs 16:47:53 Integrated Voltage Measurement... 16:48:02 Voltage Offset Correction: v_off_ini = 1009.288 uV, v_off_fin = 996.590 uV 16:48:02 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 61.718 uVs 16:48:02 Integrated Voltage Measurement... 16:48:11 Voltage Offset Correction: v_off_ini = 985.183 uV, v_off_fin = 1001.378 uV 16:48:11 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -61.630 uVs 16:48:11 Integrated Voltage Measurement... 16:48:20 Voltage Offset Correction: v_off_ini = 1001.625 uV, v_off_fin = 996.796 uV 16:48:20 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 84.107 uVs 16:48:20 Integrated Voltage Measurement... 16:48:29 Voltage Offset Correction: v_off_ini = 1002.605 uV, v_off_fin = 1015.931 uV 16:48:29 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -81.258 uVs 16:48:29 Integrated Voltage Measurement... 16:48:38 Voltage Offset Correction: v_off_ini = 991.345 uV, v_off_fin = 998.420 uV 16:48:38 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 65.269 uVs 16:48:42 I1X Average and RMS of Measurements VT = -71.090 +- 2.799 uVs I1X = 14.218 +- 0.559 uTm 16:48:45 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:48:49 Integrated Voltage Measurement... 16:48:58 Voltage Offset Correction: v_off_ini = 996.075 uV, v_off_fin = 972.790 uV 16:48:58 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -220.435 uVs 16:48:58 Integrated Voltage Measurement... 16:49:07 Voltage Offset Correction: v_off_ini = 984.102 uV, v_off_fin = 981.557 uV 16:49:07 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 263.230 uVs 16:49:07 Integrated Voltage Measurement... 16:49:16 Voltage Offset Correction: v_off_ini = 962.192 uV, v_off_fin = 965.765 uV 16:49:16 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -241.074 uVs 16:49:16 Integrated Voltage Measurement... 16:49:25 Voltage Offset Correction: v_off_ini = 981.072 uV, v_off_fin = 992.202 uV 16:49:25 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 247.843 uVs 16:49:25 Integrated Voltage Measurement... 16:49:34 Voltage Offset Correction: v_off_ini = 1000.211 uV, v_off_fin = 992.024 uV 16:49:34 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -242.042 uVs 16:49:34 Integrated Voltage Measurement... 16:49:43 Voltage Offset Correction: v_off_ini = 988.908 uV, v_off_fin = 997.134 uV 16:49:43 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 242.600 uVs 16:49:47 I1Y Average and RMS of Measurements VT = -242.871 +- 1.140 uVs I1Y = -48.574 +- 0.228 uTm 16:49:50 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:49:52 Integrated Voltage Measurement... 16:50:01 Voltage Offset Correction: v_off_ini = 996.692 uV, v_off_fin = 979.904 uV 16:50:01 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -57.182 uVs 16:50:01 Integrated Voltage Measurement... 16:50:10 Voltage Offset Correction: v_off_ini = 994.522 uV, v_off_fin = 999.742 uV 16:50:10 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 78.352 uVs 16:50:10 Integrated Voltage Measurement... 16:50:19 Voltage Offset Correction: v_off_ini = 1003.241 uV, v_off_fin = 1005.400 uV 16:50:19 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -78.543 uVs 16:50:19 Integrated Voltage Measurement... 16:50:28 Voltage Offset Correction: v_off_ini = 1006.892 uV, v_off_fin = 1002.252 uV 16:50:28 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 59.770 uVs 16:50:28 Integrated Voltage Measurement... 16:50:37 Voltage Offset Correction: v_off_ini = 1009.214 uV, v_off_fin = 1013.189 uV 16:50:37 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -99.962 uVs 16:50:37 Integrated Voltage Measurement... 16:50:46 Voltage Offset Correction: v_off_ini = 1018.260 uV, v_off_fin = 1016.061 uV 16:50:46 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 82.750 uVs 16:50:49 I2X Average and RMS of Measurements VT = -76.093 +- 10.807 uVs I2X = 7.001 +- 0.994 uTm^2 16:50:51 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 16:50:54 Integrated Voltage Measurement... 16:51:03 Voltage Offset Correction: v_off_ini = 1007.377 uV, v_off_fin = 1008.224 uV 16:51:03 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -245.362 uVs 16:51:03 Integrated Voltage Measurement... 16:51:12 Voltage Offset Correction: v_off_ini = 1015.184 uV, v_off_fin = 1005.270 uV 16:51:12 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 233.517 uVs 16:51:12 Integrated Voltage Measurement... 16:51:21 Voltage Offset Correction: v_off_ini = 1001.879 uV, v_off_fin = 1001.790 uV 16:51:21 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = -257.862 uVs 16:51:21 Integrated Voltage Measurement... 16:51:30 Voltage Offset Correction: v_off_ini = 1008.199 uV, v_off_fin = 1002.206 uV 16:51:30 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 250.031 uVs 16:51:30 Integrated Voltage Measurement... 16:51:39 Voltage Offset Correction: v_off_ini = 989.864 uV, v_off_fin = 992.569 uV 16:51:39 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -265.473 uVs 16:51:39 Integrated Voltage Measurement... 16:51:48 Voltage Offset Correction: v_off_ini = 991.041 uV, v_off_fin = 994.854 uV 16:51:48 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 259.716 uVs 16:51:51 I2Y Average and RMS of Measurements VT = -251.994 +- 9.553 uVs I2Y = -23.184 +- 0.879 uTm^2 16:51:55 Wire Position: X0 = 0.0000 mm, Y0 = 0.5000 mm 16:51:59 Integrated Voltage Measurement... 16:52:08 Voltage Offset Correction: v_off_ini = 1004.314 uV, v_off_fin = 1009.841 uV 16:52:08 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -72.645 uVs 16:52:08 Integrated Voltage Measurement... 16:52:17 Voltage Offset Correction: v_off_ini = 1007.809 uV, v_off_fin = 991.747 uV 16:52:17 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 65.601 uVs 16:52:17 Integrated Voltage Measurement... 16:52:26 Voltage Offset Correction: v_off_ini = 999.444 uV, v_off_fin = 1021.350 uV 16:52:26 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -50.741 uVs 16:52:26 Integrated Voltage Measurement... 16:52:35 Voltage Offset Correction: v_off_ini = 1014.592 uV, v_off_fin = 1006.924 uV 16:52:35 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 58.290 uVs 16:52:35 Integrated Voltage Measurement... 16:52:44 Voltage Offset Correction: v_off_ini = 996.639 uV, v_off_fin = 998.565 uV 16:52:44 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -73.831 uVs 16:52:44 Integrated Voltage Measurement... 16:52:53 Voltage Offset Correction: v_off_ini = 1010.080 uV, v_off_fin = 1005.667 uV 16:52:53 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 64.791 uVs 16:52:57 I1X Average and RMS of Measurements VT = -64.316 +- 6.931 uVs I1X = 12.864 +- 1.386 uTm 16:53:00 Wire Position: X0 = 0.0000 mm, Y0 = 0.5000 mm 16:53:04 Integrated Voltage Measurement... 16:53:13 Voltage Offset Correction: v_off_ini = 997.521 uV, v_off_fin = 997.173 uV 16:53:13 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = -209.312 uVs 16:53:13 Integrated Voltage Measurement... 16:53:22 Voltage Offset Correction: v_off_ini = 1009.254 uV, v_off_fin = 1026.339 uV 16:53:22 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 223.149 uVs 16:53:22 Integrated Voltage Measurement... 16:53:31 Voltage Offset Correction: v_off_ini = 999.426 uV, v_off_fin = 985.006 uV 16:53:31 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -247.280 uVs 16:53:31 Integrated Voltage Measurement... 16:53:40 Voltage Offset Correction: v_off_ini = 984.698 uV, v_off_fin = 1003.235 uV 16:53:40 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 244.218 uVs 16:53:40 Integrated Voltage Measurement... 16:53:49 Voltage Offset Correction: v_off_ini = 1022.171 uV, v_off_fin = 1007.813 uV 16:53:49 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -231.284 uVs 16:53:49 Integrated Voltage Measurement... 16:53:58 Voltage Offset Correction: v_off_ini = 971.988 uV, v_off_fin = 999.406 uV 16:53:58 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 257.932 uVs 16:54:02 I1Y Average and RMS of Measurements VT = -235.529 +- 13.654 uVs I1Y = -47.104 +- 2.728 uTm 16:54:05 Wire Position: X0 = 0.0000 mm, Y0 = 0.5000 mm 16:54:07 Integrated Voltage Measurement... 16:54:16 Voltage Offset Correction: v_off_ini = 988.338 uV, v_off_fin = 985.829 uV 16:54:16 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -94.251 uVs 16:54:16 Integrated Voltage Measurement... 16:54:25 Voltage Offset Correction: v_off_ini = 992.835 uV, v_off_fin = 994.077 uV 16:54:25 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 71.908 uVs 16:54:25 Integrated Voltage Measurement... 16:54:34 Voltage Offset Correction: v_off_ini = 975.387 uV, v_off_fin = 991.775 uV 16:54:34 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -80.765 uVs 16:54:34 Integrated Voltage Measurement... 16:54:43 Voltage Offset Correction: v_off_ini = 990.332 uV, v_off_fin = 986.373 uV 16:54:43 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 72.955 uVs 16:54:43 Integrated Voltage Measurement... 16:54:52 Voltage Offset Correction: v_off_ini = 982.379 uV, v_off_fin = 986.036 uV 16:54:52 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -124.262 uVs 16:54:52 Integrated Voltage Measurement... 16:55:01 Voltage Offset Correction: v_off_ini = 986.453 uV, v_off_fin = 997.459 uV 16:55:01 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 53.725 uVs 16:55:04 I2X Average and RMS of Measurements VT = -82.978 +- 4.954 uVs I2X = 7.634 +- 0.455 uTm^2 16:55:06 Wire Position: X0 = 0.0000 mm, Y0 = 0.5000 mm 16:55:09 Integrated Voltage Measurement... 16:55:18 Voltage Offset Correction: v_off_ini = 1004.056 uV, v_off_fin = 955.968 uV 16:55:18 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -227.012 uVs 16:55:18 Integrated Voltage Measurement... 16:55:27 Voltage Offset Correction: v_off_ini = 984.072 uV, v_off_fin = 992.752 uV 16:55:27 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 273.638 uVs 16:55:27 Integrated Voltage Measurement... 16:55:36 Voltage Offset Correction: v_off_ini = 988.446 uV, v_off_fin = 982.995 uV 16:55:36 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -266.496 uVs 16:55:36 Integrated Voltage Measurement... 16:55:45 Voltage Offset Correction: v_off_ini = 998.069 uV, v_off_fin = 997.388 uV 16:55:45 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 263.065 uVs 16:55:45 Integrated Voltage Measurement... 16:55:54 Voltage Offset Correction: v_off_ini = 977.588 uV, v_off_fin = 980.300 uV 16:55:54 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -261.509 uVs 16:55:54 Integrated Voltage Measurement... 16:56:03 Voltage Offset Correction: v_off_ini = 973.234 uV, v_off_fin = 974.383 uV 16:56:03 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 295.637 uVs 16:56:06 I2Y Average and RMS of Measurements VT = -264.559 +- 11.533 uVs I2Y = -24.339 +- 1.061 uTm^2 16:56:10 Wire Position: X0 = 0.0000 mm, Y0 = 1.0000 mm 16:56:14 Integrated Voltage Measurement... 16:56:23 Voltage Offset Correction: v_off_ini = 993.244 uV, v_off_fin = 964.800 uV 16:56:23 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.4999 mm dXY = -0.4999 mm VT = -68.416 uVs 16:56:23 Integrated Voltage Measurement... 16:56:32 Voltage Offset Correction: v_off_ini = 973.385 uV, v_off_fin = 995.086 uV 16:56:32 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 39.130 uVs 16:56:32 Integrated Voltage Measurement... 16:56:41 Voltage Offset Correction: v_off_ini = 980.725 uV, v_off_fin = 979.943 uV 16:56:41 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -58.706 uVs 16:56:41 Integrated Voltage Measurement... 16:56:50 Voltage Offset Correction: v_off_ini = 971.765 uV, v_off_fin = 976.471 uV 16:56:50 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 32.519 uVs 16:56:50 Integrated Voltage Measurement... 16:56:59 Voltage Offset Correction: v_off_ini = 974.130 uV, v_off_fin = 971.947 uV 16:56:59 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -61.648 uVs 16:56:59 Integrated Voltage Measurement... 16:57:08 Voltage Offset Correction: v_off_ini = 973.441 uV, v_off_fin = 984.479 uV 16:57:08 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 57.365 uVs 16:57:12 I1X Average and RMS of Measurements VT = -52.964 +- 5.701 uVs I1X = 10.593 +- 1.140 uTm 16:57:15 Wire Position: X0 = 0.0000 mm, Y0 = 1.0000 mm 16:57:19 Integrated Voltage Measurement... 16:57:28 Voltage Offset Correction: v_off_ini = 962.729 uV, v_off_fin = 949.047 uV 16:57:28 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = -234.199 uVs 16:57:28 Integrated Voltage Measurement... 16:57:37 Voltage Offset Correction: v_off_ini = 961.743 uV, v_off_fin = 987.969 uV 16:57:37 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 226.639 uVs 16:57:37 Integrated Voltage Measurement... 16:57:46 Voltage Offset Correction: v_off_ini = 975.787 uV, v_off_fin = 964.295 uV 16:57:46 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -234.753 uVs 16:57:46 Integrated Voltage Measurement... 16:57:55 Voltage Offset Correction: v_off_ini = 980.835 uV, v_off_fin = 981.076 uV 16:57:55 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 234.996 uVs 16:57:55 Integrated Voltage Measurement... 16:58:04 Voltage Offset Correction: v_off_ini = 962.113 uV, v_off_fin = 959.510 uV 16:58:04 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -263.559 uVs 16:58:04 Integrated Voltage Measurement... 16:58:13 Voltage Offset Correction: v_off_ini = 977.284 uV, v_off_fin = 993.579 uV 16:58:13 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 225.314 uVs 16:58:17 I1Y Average and RMS of Measurements VT = -236.577 +- 5.848 uVs I1Y = -47.314 +- 1.167 uTm 16:58:20 Wire Position: X0 = 0.0000 mm, Y0 = 1.0000 mm 16:58:22 Integrated Voltage Measurement... 16:58:31 Voltage Offset Correction: v_off_ini = 966.673 uV, v_off_fin = 979.536 uV 16:58:32 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -86.563 uVs 16:58:32 Integrated Voltage Measurement... 16:58:41 Voltage Offset Correction: v_off_ini = 984.286 uV, v_off_fin = 967.227 uV 16:58:41 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 76.342 uVs 16:58:41 Integrated Voltage Measurement... 16:58:50 Voltage Offset Correction: v_off_ini = 977.015 uV, v_off_fin = 973.107 uV 16:58:50 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -92.116 uVs 16:58:50 Integrated Voltage Measurement... 16:58:59 Voltage Offset Correction: v_off_ini = 987.194 uV, v_off_fin = 980.242 uV 16:58:59 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 81.514 uVs 16:58:59 Integrated Voltage Measurement... 16:59:08 Voltage Offset Correction: v_off_ini = 976.552 uV, v_off_fin = 989.755 uV 16:59:08 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -72.977 uVs 16:59:08 Integrated Voltage Measurement... 16:59:17 Voltage Offset Correction: v_off_ini = 999.201 uV, v_off_fin = 1007.870 uV 16:59:17 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 55.925 uVs 16:59:19 I2X Average and RMS of Measurements VT = -77.573 +- 9.533 uVs I2X = 7.137 +- 0.877 uTm^2 16:59:22 Wire Position: X0 = 0.0000 mm, Y0 = 1.0000 mm 16:59:24 Integrated Voltage Measurement... 16:59:33 Voltage Offset Correction: v_off_ini = 990.833 uV, v_off_fin = 984.708 uV 16:59:33 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -252.135 uVs 16:59:33 Integrated Voltage Measurement... 16:59:42 Voltage Offset Correction: v_off_ini = 988.411 uV, v_off_fin = 984.118 uV 16:59:42 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 275.231 uVs 16:59:42 Integrated Voltage Measurement... 16:59:51 Voltage Offset Correction: v_off_ini = 995.516 uV, v_off_fin = 989.978 uV 16:59:51 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -249.890 uVs 16:59:51 Integrated Voltage Measurement... 17:00:00 Voltage Offset Correction: v_off_ini = 965.029 uV, v_off_fin = 970.801 uV 17:00:00 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 279.032 uVs 17:00:00 Integrated Voltage Measurement... 17:00:09 Voltage Offset Correction: v_off_ini = 969.640 uV, v_off_fin = 978.834 uV 17:00:10 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -240.512 uVs 17:00:10 Integrated Voltage Measurement... 17:00:19 Voltage Offset Correction: v_off_ini = 966.897 uV, v_off_fin = 977.986 uV 17:00:19 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 267.378 uVs 17:00:21 I2Y Average and RMS of Measurements VT = -260.696 +- 4.785 uVs I2Y = -23.984 +- 0.440 uTm^2 17:00:26 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm