SLAC Magnetic Measurements Date: 01-07-2025 Time: 09:38:38 Project: LCLS-II-HE Magnet Type: Phase Shifter Magnet Name: HE_SXPS_10086 Dataset Number: 1 Run Purpose: Tuning Run Type: Stretched Wire Run Number: 4 Measurement Device: On Axis Operator: YL Comment: 0.375mm to 0.400mm 09:38:45 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:38:48 Motion Parameters, X Controller, Axis 1 acc = 0.001000 m/s/s dcl = 0.001000 m/s/s vel = 0.001000 m/s pos = 0.000000 m kp = 188.600000 ki = 820.000000 kd = 491.400000 db = 0.000000 09:38:49 Motion Parameters, X Controller, Axis 2 acc = 0.001000 m/s/s dcl = 0.001000 m/s/s vel = 0.001000 m/s pos = 0.000000 m kp = 188.600000 ki = 820.000000 kd = 491.400000 db = 0.000000 09:38:51 Motion Parameters, Y Controller, Axis 1 acc = 0.001000 m/s/s dcl = 0.001000 m/s/s vel = 0.001000 m/s pos = 0.000000 m kp = 188.600000 ki = 820.000000 kd = 491.400000 db = 0.000000 09:38:52 Motion Parameters, Y Controller, Axis 2 acc = 0.001000 m/s/s dcl = 0.001000 m/s/s vel = 0.001000 m/s pos = 0.000000 m kp = 188.600000 ki = 820.000000 kd = 491.400000 db = 0.000000 09:39:24 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:39:28 Integrated Voltage Measurement... 09:39:37 Voltage Offset Correction: v_off_ini = -92.749 uV, v_off_fin = -88.760 uV 09:39:37 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.4998 mm dXY = -0.4999 mm VT = -112.748 uVs 09:39:37 Integrated Voltage Measurement... 09:39:46 Voltage Offset Correction: v_off_ini = -85.153 uV, v_off_fin = -76.709 uV 09:39:46 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = 107.430 uVs 09:39:46 Integrated Voltage Measurement... 09:39:55 Voltage Offset Correction: v_off_ini = -49.511 uV, v_off_fin = -54.317 uV 09:39:55 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -103.889 uVs 09:39:55 Integrated Voltage Measurement... 09:40:04 Voltage Offset Correction: v_off_ini = -50.096 uV, v_off_fin = -48.344 uV 09:40:04 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 120.806 uVs 09:40:04 Integrated Voltage Measurement... 09:40:13 Voltage Offset Correction: v_off_ini = -46.040 uV, v_off_fin = -45.991 uV 09:40:13 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -107.680 uVs 09:40:13 Integrated Voltage Measurement... 09:40:22 Voltage Offset Correction: v_off_ini = -55.196 uV, v_off_fin = -55.388 uV 09:40:22 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 109.843 uVs 09:40:27 I1X Average and RMS of Measurements VT = -110.399 +- 1.480 uVs I1X = 22.081 +- 0.296 uTm 09:40:29 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:40:33 Integrated Voltage Measurement... 09:40:42 Voltage Offset Correction: v_off_ini = -53.512 uV, v_off_fin = -54.064 uV 09:40:42 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = -188.051 uVs 09:40:42 Integrated Voltage Measurement... 09:40:51 Voltage Offset Correction: v_off_ini = -67.063 uV, v_off_fin = -50.393 uV 09:40:51 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 185.618 uVs 09:40:51 Integrated Voltage Measurement... 09:41:00 Voltage Offset Correction: v_off_ini = -49.925 uV, v_off_fin = -56.644 uV 09:41:00 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -193.841 uVs 09:41:00 Integrated Voltage Measurement... 09:41:09 Voltage Offset Correction: v_off_ini = -48.515 uV, v_off_fin = -29.966 uV 09:41:09 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 197.800 uVs 09:41:09 Integrated Voltage Measurement... 09:41:18 Voltage Offset Correction: v_off_ini = -34.173 uV, v_off_fin = -40.010 uV 09:41:18 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -187.214 uVs 09:41:18 Integrated Voltage Measurement... 09:41:27 Voltage Offset Correction: v_off_ini = -33.205 uV, v_off_fin = -18.291 uV 09:41:27 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 192.466 uVs 09:41:32 I1Y Average and RMS of Measurements VT = -190.832 +- 3.735 uVs I1Y = -38.167 +- 0.746 uTm 09:41:37 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:41:39 Integrated Voltage Measurement... 09:41:48 Voltage Offset Correction: v_off_ini = -30.319 uV, v_off_fin = -42.935 uV 09:41:48 Integrated Voltage Measurement dXYus = -1.0002 mm, dXYds = 0.0000 mm dXY = -1.0002 mm VT = -121.879 uVs 09:41:48 Integrated Voltage Measurement... 09:41:57 Voltage Offset Correction: v_off_ini = -18.934 uV, v_off_fin = 9.253 uV 09:41:57 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 126.831 uVs 09:41:57 Integrated Voltage Measurement... 09:42:06 Voltage Offset Correction: v_off_ini = -16.474 uV, v_off_fin = -15.660 uV 09:42:06 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -120.758 uVs 09:42:06 Integrated Voltage Measurement... 09:42:15 Voltage Offset Correction: v_off_ini = -0.720 uV, v_off_fin = -2.349 uV 09:42:15 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 100.160 uVs 09:42:15 Integrated Voltage Measurement... 09:42:24 Voltage Offset Correction: v_off_ini = -8.366 uV, v_off_fin = -8.613 uV 09:42:24 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -106.420 uVs 09:42:25 Integrated Voltage Measurement... 09:42:33 Voltage Offset Correction: v_off_ini = 16.459 uV, v_off_fin = 20.987 uV 09:42:34 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 123.472 uVs 09:42:36 I2X Average and RMS of Measurements VT = -116.587 +- 5.790 uVs I2X = 10.726 +- 0.532 uTm^2 09:42:41 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:42:44 Integrated Voltage Measurement... 09:42:53 Voltage Offset Correction: v_off_ini = 27.040 uV, v_off_fin = 12.321 uV 09:42:53 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -295.662 uVs 09:42:53 Integrated Voltage Measurement... 09:43:02 Voltage Offset Correction: v_off_ini = 30.438 uV, v_off_fin = 32.070 uV 09:43:02 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 278.651 uVs 09:43:02 Integrated Voltage Measurement... 09:43:11 Voltage Offset Correction: v_off_ini = 45.301 uV, v_off_fin = 48.314 uV 09:43:11 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -258.506 uVs 09:43:11 Integrated Voltage Measurement... 09:43:20 Voltage Offset Correction: v_off_ini = 62.426 uV, v_off_fin = 62.655 uV 09:43:20 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 269.298 uVs 09:43:20 Integrated Voltage Measurement... 09:43:29 Voltage Offset Correction: v_off_ini = 49.622 uV, v_off_fin = 51.326 uV 09:43:29 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -272.302 uVs 09:43:29 Integrated Voltage Measurement... 09:43:38 Voltage Offset Correction: v_off_ini = 64.428 uV, v_off_fin = 49.537 uV 09:43:38 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 259.939 uVs 09:43:40 I2Y Average and RMS of Measurements VT = -272.393 +- 10.479 uVs I2Y = -25.060 +- 0.964 uTm^2 09:43:50 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:43:54 Integrated Voltage Measurement... 09:44:03 Voltage Offset Correction: v_off_ini = 68.253 uV, v_off_fin = 43.438 uV 09:44:03 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.4998 mm dXY = -0.4999 mm VT = -93.675 uVs 09:44:03 Integrated Voltage Measurement... 09:44:12 Voltage Offset Correction: v_off_ini = 65.021 uV, v_off_fin = 50.001 uV 09:44:12 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 93.250 uVs 09:44:12 Integrated Voltage Measurement... 09:44:21 Voltage Offset Correction: v_off_ini = 68.896 uV, v_off_fin = 67.325 uV 09:44:21 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -92.671 uVs 09:44:21 Integrated Voltage Measurement... 09:44:30 Voltage Offset Correction: v_off_ini = 51.828 uV, v_off_fin = 39.482 uV 09:44:30 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 110.007 uVs 09:44:30 Integrated Voltage Measurement... 09:44:39 Voltage Offset Correction: v_off_ini = 40.653 uV, v_off_fin = 33.298 uV 09:44:39 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -164.102 uVs 09:44:39 Integrated Voltage Measurement... 09:44:48 Voltage Offset Correction: v_off_ini = 33.735 uV, v_off_fin = 43.349 uV 09:44:48 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 107.330 uVs 09:44:53 I1X Average and RMS of Measurements VT = -110.173 +- 18.346 uVs I1X = 22.035 +- 3.668 uTm 09:44:55 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:44:59 Integrated Voltage Measurement... 09:45:08 Voltage Offset Correction: v_off_ini = 40.989 uV, v_off_fin = 47.291 uV 09:45:08 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.4999 mm dXY = 0.4999 mm VT = -182.211 uVs 09:45:08 Integrated Voltage Measurement... 09:45:17 Voltage Offset Correction: v_off_ini = 55.657 uV, v_off_fin = 62.052 uV 09:45:17 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 193.283 uVs 09:45:17 Integrated Voltage Measurement... 09:45:26 Voltage Offset Correction: v_off_ini = 57.966 uV, v_off_fin = 47.916 uV 09:45:26 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -176.468 uVs 09:45:26 Integrated Voltage Measurement... 09:45:35 Voltage Offset Correction: v_off_ini = 60.970 uV, v_off_fin = 83.368 uV 09:45:35 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 195.366 uVs 09:45:35 Integrated Voltage Measurement... 09:45:44 Voltage Offset Correction: v_off_ini = 64.645 uV, v_off_fin = 57.402 uV 09:45:44 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -203.438 uVs 09:45:44 Integrated Voltage Measurement... 09:45:53 Voltage Offset Correction: v_off_ini = 70.614 uV, v_off_fin = 81.672 uV 09:45:53 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 197.745 uVs 09:45:58 I1Y Average and RMS of Measurements VT = -191.418 +- 6.529 uVs I1Y = -38.285 +- 1.305 uTm 09:46:03 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:46:05 Integrated Voltage Measurement... 09:46:14 Voltage Offset Correction: v_off_ini = 38.049 uV, v_off_fin = 47.585 uV 09:46:14 Integrated Voltage Measurement dXYus = -1.0002 mm, dXYds = 0.0000 mm dXY = -1.0002 mm VT = -90.514 uVs 09:46:14 Integrated Voltage Measurement... 09:46:23 Voltage Offset Correction: v_off_ini = 56.401 uV, v_off_fin = 64.662 uV 09:46:23 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 114.548 uVs 09:46:23 Integrated Voltage Measurement... 09:46:32 Voltage Offset Correction: v_off_ini = 56.520 uV, v_off_fin = 64.828 uV 09:46:32 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -90.827 uVs 09:46:32 Integrated Voltage Measurement... 09:46:41 Voltage Offset Correction: v_off_ini = 85.975 uV, v_off_fin = 103.845 uV 09:46:42 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 105.348 uVs 09:46:42 Integrated Voltage Measurement... 09:46:51 Voltage Offset Correction: v_off_ini = 73.079 uV, v_off_fin = 73.774 uV 09:46:51 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -109.389 uVs 09:46:51 Integrated Voltage Measurement... 09:47:00 Voltage Offset Correction: v_off_ini = 95.896 uV, v_off_fin = 98.452 uV 09:47:00 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 71.385 uVs 09:47:02 I2X Average and RMS of Measurements VT = -97.002 +- 5.017 uVs I2X = 8.924 +- 0.461 uTm^2 09:47:07 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:47:10 Integrated Voltage Measurement... 09:47:19 Voltage Offset Correction: v_off_ini = 92.994 uV, v_off_fin = 77.765 uV 09:47:19 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -267.733 uVs 09:47:19 Integrated Voltage Measurement... 09:47:28 Voltage Offset Correction: v_off_ini = 94.022 uV, v_off_fin = 104.165 uV 09:47:28 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 265.222 uVs 09:47:28 Integrated Voltage Measurement... 09:47:37 Voltage Offset Correction: v_off_ini = 82.903 uV, v_off_fin = 91.667 uV 09:47:37 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -267.274 uVs 09:47:37 Integrated Voltage Measurement... 09:47:46 Voltage Offset Correction: v_off_ini = 104.462 uV, v_off_fin = 100.682 uV 09:47:46 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 265.246 uVs 09:47:46 Integrated Voltage Measurement... 09:47:55 Voltage Offset Correction: v_off_ini = 80.266 uV, v_off_fin = 74.671 uV 09:47:55 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -262.639 uVs 09:47:55 Integrated Voltage Measurement... 09:48:04 Voltage Offset Correction: v_off_ini = 88.435 uV, v_off_fin = 90.660 uV 09:48:04 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 280.380 uVs 09:48:06 I2Y Average and RMS of Measurements VT = -268.082 +- 2.425 uVs I2Y = -24.664 +- 0.223 uTm^2 09:48:16 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:48:20 Integrated Voltage Measurement... 09:48:29 Voltage Offset Correction: v_off_ini = 95.687 uV, v_off_fin = 89.860 uV 09:48:29 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -88.096 uVs 09:48:29 Integrated Voltage Measurement... 09:48:38 Voltage Offset Correction: v_off_ini = 87.187 uV, v_off_fin = 82.556 uV 09:48:38 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = 108.793 uVs 09:48:38 Integrated Voltage Measurement... 09:48:47 Voltage Offset Correction: v_off_ini = 99.194 uV, v_off_fin = 84.621 uV 09:48:47 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -105.255 uVs 09:48:47 Integrated Voltage Measurement... 09:48:56 Voltage Offset Correction: v_off_ini = 95.285 uV, v_off_fin = 86.660 uV 09:48:56 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 90.942 uVs 09:48:56 Integrated Voltage Measurement... 09:49:05 Voltage Offset Correction: v_off_ini = 88.538 uV, v_off_fin = 82.706 uV 09:49:05 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -97.500 uVs 09:49:05 Integrated Voltage Measurement... 09:49:14 Voltage Offset Correction: v_off_ini = 67.784 uV, v_off_fin = 100.928 uV 09:49:14 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 118.633 uVs 09:49:19 I1X Average and RMS of Measurements VT = -101.537 +- 4.620 uVs I1X = 20.307 +- 0.924 uTm 09:49:21 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:49:25 Integrated Voltage Measurement... 09:49:34 Voltage Offset Correction: v_off_ini = 98.149 uV, v_off_fin = 102.672 uV 09:49:34 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = -208.464 uVs 09:49:34 Integrated Voltage Measurement... 09:49:43 Voltage Offset Correction: v_off_ini = 108.835 uV, v_off_fin = 123.065 uV 09:49:43 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 196.730 uVs 09:49:43 Integrated Voltage Measurement... 09:49:52 Voltage Offset Correction: v_off_ini = 105.303 uV, v_off_fin = 76.596 uV 09:49:52 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -178.994 uVs 09:49:52 Integrated Voltage Measurement... 09:50:01 Voltage Offset Correction: v_off_ini = 105.782 uV, v_off_fin = 112.666 uV 09:50:01 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 177.568 uVs 09:50:01 Integrated Voltage Measurement... 09:50:10 Voltage Offset Correction: v_off_ini = 104.130 uV, v_off_fin = 93.063 uV 09:50:10 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -217.020 uVs 09:50:10 Integrated Voltage Measurement... 09:50:19 Voltage Offset Correction: v_off_ini = 112.413 uV, v_off_fin = 112.275 uV 09:50:19 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 174.559 uVs 09:50:24 I1Y Average and RMS of Measurements VT = -192.223 +- 10.242 uVs I1Y = -38.445 +- 2.049 uTm 09:50:29 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:50:31 Integrated Voltage Measurement... 09:50:40 Voltage Offset Correction: v_off_ini = 84.849 uV, v_off_fin = 84.088 uV 09:50:40 Integrated Voltage Measurement dXYus = -1.0002 mm, dXYds = 0.0000 mm dXY = -1.0002 mm VT = -111.986 uVs 09:50:40 Integrated Voltage Measurement... 09:50:49 Voltage Offset Correction: v_off_ini = 95.078 uV, v_off_fin = 106.263 uV 09:50:49 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 100.594 uVs 09:50:49 Integrated Voltage Measurement... 09:50:58 Voltage Offset Correction: v_off_ini = 119.413 uV, v_off_fin = 110.816 uV 09:50:58 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -96.138 uVs 09:50:58 Integrated Voltage Measurement... 09:51:07 Voltage Offset Correction: v_off_ini = 97.606 uV, v_off_fin = 100.823 uV 09:51:07 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 91.490 uVs 09:51:08 Integrated Voltage Measurement... 09:51:16 Voltage Offset Correction: v_off_ini = 117.423 uV, v_off_fin = 113.020 uV 09:51:17 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -113.165 uVs 09:51:17 Integrated Voltage Measurement... 09:51:26 Voltage Offset Correction: v_off_ini = 105.221 uV, v_off_fin = 101.202 uV 09:51:26 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 115.583 uVs 09:51:28 I2X Average and RMS of Measurements VT = -104.826 +- 8.457 uVs I2X = 9.644 +- 0.778 uTm^2 09:51:33 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:51:36 Integrated Voltage Measurement... 09:51:45 Voltage Offset Correction: v_off_ini = 124.973 uV, v_off_fin = 116.650 uV 09:51:45 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -285.955 uVs 09:51:45 Integrated Voltage Measurement... 09:51:54 Voltage Offset Correction: v_off_ini = 108.842 uV, v_off_fin = 93.827 uV 09:51:54 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 265.703 uVs 09:51:54 Integrated Voltage Measurement... 09:52:03 Voltage Offset Correction: v_off_ini = 98.960 uV, v_off_fin = 114.021 uV 09:52:03 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -245.149 uVs 09:52:03 Integrated Voltage Measurement... 09:52:12 Voltage Offset Correction: v_off_ini = 111.793 uV, v_off_fin = 99.950 uV 09:52:12 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 257.942 uVs 09:52:12 Integrated Voltage Measurement... 09:52:21 Voltage Offset Correction: v_off_ini = 97.688 uV, v_off_fin = 112.855 uV 09:52:21 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -251.670 uVs 09:52:21 Integrated Voltage Measurement... 09:52:30 Voltage Offset Correction: v_off_ini = 106.091 uV, v_off_fin = 99.149 uV 09:52:30 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 275.128 uVs 09:52:32 I2Y Average and RMS of Measurements VT = -263.591 +- 9.915 uVs I2Y = -24.250 +- 0.912 uTm^2 09:52:42 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:52:46 Integrated Voltage Measurement... 09:52:55 Voltage Offset Correction: v_off_ini = 103.103 uV, v_off_fin = 102.303 uV 09:52:55 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -100.005 uVs 09:52:55 Integrated Voltage Measurement... 09:53:04 Voltage Offset Correction: v_off_ini = 102.127 uV, v_off_fin = 101.591 uV 09:53:04 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 111.909 uVs 09:53:04 Integrated Voltage Measurement... 09:53:13 Voltage Offset Correction: v_off_ini = 111.748 uV, v_off_fin = 112.236 uV 09:53:13 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -110.112 uVs 09:53:13 Integrated Voltage Measurement... 09:53:22 Voltage Offset Correction: v_off_ini = 108.768 uV, v_off_fin = 118.718 uV 09:53:22 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 113.734 uVs 09:53:22 Integrated Voltage Measurement... 09:53:31 Voltage Offset Correction: v_off_ini = 133.723 uV, v_off_fin = 121.071 uV 09:53:31 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -101.521 uVs 09:53:31 Integrated Voltage Measurement... 09:53:40 Voltage Offset Correction: v_off_ini = 104.342 uV, v_off_fin = 87.831 uV 09:53:40 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 107.314 uVs 09:53:45 I1X Average and RMS of Measurements VT = -107.433 +- 3.237 uVs I1X = 21.487 +- 0.647 uTm 09:53:47 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:53:51 Integrated Voltage Measurement... 09:54:00 Voltage Offset Correction: v_off_ini = 101.590 uV, v_off_fin = 83.553 uV 09:54:00 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -207.870 uVs 09:54:00 Integrated Voltage Measurement... 09:54:09 Voltage Offset Correction: v_off_ini = 102.818 uV, v_off_fin = 113.884 uV 09:54:10 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 189.263 uVs 09:54:10 Integrated Voltage Measurement... 09:54:19 Voltage Offset Correction: v_off_ini = 107.111 uV, v_off_fin = 99.841 uV 09:54:19 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -193.250 uVs 09:54:19 Integrated Voltage Measurement... 09:54:28 Voltage Offset Correction: v_off_ini = 103.820 uV, v_off_fin = 125.852 uV 09:54:28 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 197.934 uVs 09:54:28 Integrated Voltage Measurement... 09:54:37 Voltage Offset Correction: v_off_ini = 122.747 uV, v_off_fin = 106.238 uV 09:54:37 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -194.119 uVs 09:54:37 Integrated Voltage Measurement... 09:54:46 Voltage Offset Correction: v_off_ini = 114.753 uV, v_off_fin = 121.578 uV 09:54:46 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 174.432 uVs 09:54:50 I1Y Average and RMS of Measurements VT = -192.811 +- 6.157 uVs I1Y = -38.560 +- 1.231 uTm 09:54:55 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:54:58 Integrated Voltage Measurement... 09:55:07 Voltage Offset Correction: v_off_ini = 114.551 uV, v_off_fin = 108.081 uV 09:55:07 Integrated Voltage Measurement dXYus = -1.0002 mm, dXYds = 0.0000 mm dXY = -1.0002 mm VT = -91.715 uVs 09:55:07 Integrated Voltage Measurement... 09:55:16 Voltage Offset Correction: v_off_ini = 109.862 uV, v_off_fin = 136.891 uV 09:55:16 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 113.497 uVs 09:55:16 Integrated Voltage Measurement... 09:55:25 Voltage Offset Correction: v_off_ini = 145.694 uV, v_off_fin = 140.286 uV 09:55:25 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -96.285 uVs 09:55:25 Integrated Voltage Measurement... 09:55:34 Voltage Offset Correction: v_off_ini = 160.261 uV, v_off_fin = 162.620 uV 09:55:34 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 112.473 uVs 09:55:34 Integrated Voltage Measurement... 09:55:43 Voltage Offset Correction: v_off_ini = 144.296 uV, v_off_fin = 139.817 uV 09:55:43 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -95.694 uVs 09:55:43 Integrated Voltage Measurement... 09:55:52 Voltage Offset Correction: v_off_ini = 160.592 uV, v_off_fin = 156.968 uV 09:55:52 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 89.886 uVs 09:55:54 I2X Average and RMS of Measurements VT = -99.925 +- 5.097 uVs I2X = 9.193 +- 0.469 uTm^2 09:55:59 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:56:02 Integrated Voltage Measurement... 09:56:11 Voltage Offset Correction: v_off_ini = 128.778 uV, v_off_fin = 132.567 uV 09:56:11 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -255.254 uVs 09:56:11 Integrated Voltage Measurement... 09:56:20 Voltage Offset Correction: v_off_ini = 122.169 uV, v_off_fin = 134.317 uV 09:56:20 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 255.236 uVs 09:56:20 Integrated Voltage Measurement... 09:56:29 Voltage Offset Correction: v_off_ini = 139.510 uV, v_off_fin = 153.919 uV 09:56:29 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -235.129 uVs 09:56:29 Integrated Voltage Measurement... 09:56:38 Voltage Offset Correction: v_off_ini = 137.738 uV, v_off_fin = 125.946 uV 09:56:38 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 276.304 uVs 09:56:38 Integrated Voltage Measurement... 09:56:47 Voltage Offset Correction: v_off_ini = 153.364 uV, v_off_fin = 152.896 uV 09:56:47 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -254.545 uVs 09:56:47 Integrated Voltage Measurement... 09:56:56 Voltage Offset Correction: v_off_ini = 134.700 uV, v_off_fin = 145.465 uV 09:56:56 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 255.922 uVs 09:56:59 I2Y Average and RMS of Measurements VT = -255.398 +- 0.225 uVs I2Y = -23.497 +- 0.021 uTm^2 09:57:08 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:57:13 Integrated Voltage Measurement... 09:57:22 Voltage Offset Correction: v_off_ini = 138.350 uV, v_off_fin = 117.448 uV 09:57:22 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -92.305 uVs 09:57:22 Integrated Voltage Measurement... 09:57:31 Voltage Offset Correction: v_off_ini = 133.840 uV, v_off_fin = 118.565 uV 09:57:31 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 86.936 uVs 09:57:31 Integrated Voltage Measurement... 09:57:40 Voltage Offset Correction: v_off_ini = 120.932 uV, v_off_fin = 137.086 uV 09:57:40 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -83.616 uVs 09:57:40 Integrated Voltage Measurement... 09:57:49 Voltage Offset Correction: v_off_ini = 140.293 uV, v_off_fin = 147.538 uV 09:57:49 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 83.791 uVs 09:57:49 Integrated Voltage Measurement... 09:57:58 Voltage Offset Correction: v_off_ini = 143.934 uV, v_off_fin = 138.750 uV 09:57:58 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -95.719 uVs 09:57:58 Integrated Voltage Measurement... 09:58:07 Voltage Offset Correction: v_off_ini = 149.403 uV, v_off_fin = 126.087 uV 09:58:07 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 104.934 uVs 09:58:11 I1X Average and RMS of Measurements VT = -91.217 +- 6.880 uVs I1X = 18.244 +- 1.376 uTm 09:58:13 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:58:18 Integrated Voltage Measurement... 09:58:27 Voltage Offset Correction: v_off_ini = 153.017 uV, v_off_fin = 145.925 uV 09:58:27 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -187.659 uVs 09:58:27 Integrated Voltage Measurement... 09:58:36 Voltage Offset Correction: v_off_ini = 116.258 uV, v_off_fin = 140.735 uV 09:58:36 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 162.279 uVs 09:58:36 Integrated Voltage Measurement... 09:58:45 Voltage Offset Correction: v_off_ini = 147.872 uV, v_off_fin = 140.407 uV 09:58:45 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -206.015 uVs 09:58:45 Integrated Voltage Measurement... 09:58:54 Voltage Offset Correction: v_off_ini = 132.580 uV, v_off_fin = 147.027 uV 09:58:54 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 178.645 uVs 09:58:54 Integrated Voltage Measurement... 09:59:03 Voltage Offset Correction: v_off_ini = 146.486 uV, v_off_fin = 121.724 uV 09:59:03 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -173.872 uVs 09:59:03 Integrated Voltage Measurement... 09:59:12 Voltage Offset Correction: v_off_ini = 116.025 uV, v_off_fin = 154.031 uV 09:59:12 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 176.217 uVs 09:59:16 I1Y Average and RMS of Measurements VT = -180.781 +- 8.166 uVs I1Y = -36.154 +- 1.634 uTm 09:59:21 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 09:59:24 Integrated Voltage Measurement... 09:59:33 Voltage Offset Correction: v_off_ini = 141.746 uV, v_off_fin = 116.919 uV 09:59:33 Integrated Voltage Measurement dXYus = -1.0002 mm, dXYds = 0.0000 mm dXY = -1.0002 mm VT = -99.646 uVs 09:59:33 Integrated Voltage Measurement... 09:59:42 Voltage Offset Correction: v_off_ini = 130.558 uV, v_off_fin = 152.387 uV 09:59:42 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 90.845 uVs 09:59:42 Integrated Voltage Measurement... 09:59:51 Voltage Offset Correction: v_off_ini = 142.879 uV, v_off_fin = 135.858 uV 09:59:51 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -76.749 uVs 09:59:51 Integrated Voltage Measurement... 10:00:00 Voltage Offset Correction: v_off_ini = 144.741 uV, v_off_fin = 153.215 uV 10:00:00 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 111.854 uVs 10:00:00 Integrated Voltage Measurement... 10:00:09 Voltage Offset Correction: v_off_ini = 156.511 uV, v_off_fin = 142.875 uV 10:00:09 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -93.167 uVs 10:00:09 Integrated Voltage Measurement... 10:00:18 Voltage Offset Correction: v_off_ini = 148.068 uV, v_off_fin = 170.099 uV 10:00:18 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 116.725 uVs 10:00:20 I2X Average and RMS of Measurements VT = -98.164 +- 4.811 uVs I2X = 9.031 +- 0.443 uTm^2 10:00:25 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:00:28 Integrated Voltage Measurement... 10:00:37 Voltage Offset Correction: v_off_ini = 151.268 uV, v_off_fin = 126.151 uV 10:00:37 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -268.652 uVs 10:00:37 Integrated Voltage Measurement... 10:00:46 Voltage Offset Correction: v_off_ini = 140.601 uV, v_off_fin = 153.686 uV 10:00:46 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 266.140 uVs 10:00:46 Integrated Voltage Measurement... 10:00:55 Voltage Offset Correction: v_off_ini = 144.781 uV, v_off_fin = 147.820 uV 10:00:55 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -240.276 uVs 10:00:55 Integrated Voltage Measurement... 10:01:04 Voltage Offset Correction: v_off_ini = 145.829 uV, v_off_fin = 151.369 uV 10:01:04 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 272.781 uVs 10:01:04 Integrated Voltage Measurement... 10:01:13 Voltage Offset Correction: v_off_ini = 157.742 uV, v_off_fin = 158.764 uV 10:01:13 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -282.822 uVs 10:01:13 Integrated Voltage Measurement... 10:01:22 Voltage Offset Correction: v_off_ini = 154.516 uV, v_off_fin = 165.770 uV 10:01:22 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 269.017 uVs 10:01:24 I2Y Average and RMS of Measurements VT = -266.615 +- 7.936 uVs I2Y = -24.529 +- 0.730 uTm^2 10:01:34 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:01:38 Integrated Voltage Measurement... 10:01:47 Voltage Offset Correction: v_off_ini = 166.901 uV, v_off_fin = 177.698 uV 10:01:47 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -107.989 uVs 10:01:48 Integrated Voltage Measurement... 10:01:56 Voltage Offset Correction: v_off_ini = 164.819 uV, v_off_fin = 176.981 uV 10:01:56 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 107.925 uVs 10:01:57 Integrated Voltage Measurement... 10:02:05 Voltage Offset Correction: v_off_ini = 168.963 uV, v_off_fin = 150.713 uV 10:02:06 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -100.138 uVs 10:02:06 Integrated Voltage Measurement... 10:02:14 Voltage Offset Correction: v_off_ini = 153.244 uV, v_off_fin = 160.995 uV 10:02:15 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 85.423 uVs 10:02:15 Integrated Voltage Measurement... 10:02:24 Voltage Offset Correction: v_off_ini = 146.128 uV, v_off_fin = 148.928 uV 10:02:24 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -91.446 uVs 10:02:24 Integrated Voltage Measurement... 10:02:33 Voltage Offset Correction: v_off_ini = 161.961 uV, v_off_fin = 163.384 uV 10:02:33 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 72.339 uVs 10:02:37 I1X Average and RMS of Measurements VT = -94.210 +- 10.689 uVs I1X = 18.842 +- 2.138 uTm 10:02:39 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:02:44 Integrated Voltage Measurement... 10:02:53 Voltage Offset Correction: v_off_ini = 160.709 uV, v_off_fin = 142.170 uV 10:02:53 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = -186.638 uVs 10:02:53 Integrated Voltage Measurement... 10:03:02 Voltage Offset Correction: v_off_ini = 148.033 uV, v_off_fin = 157.775 uV 10:03:02 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 180.349 uVs 10:03:02 Integrated Voltage Measurement... 10:03:11 Voltage Offset Correction: v_off_ini = 151.380 uV, v_off_fin = 152.293 uV 10:03:11 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -186.212 uVs 10:03:11 Integrated Voltage Measurement... 10:03:20 Voltage Offset Correction: v_off_ini = 171.443 uV, v_off_fin = 181.313 uV 10:03:20 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 188.425 uVs 10:03:20 Integrated Voltage Measurement... 10:03:29 Voltage Offset Correction: v_off_ini = 173.664 uV, v_off_fin = 152.579 uV 10:03:29 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -184.201 uVs 10:03:29 Integrated Voltage Measurement... 10:03:38 Voltage Offset Correction: v_off_ini = 160.645 uV, v_off_fin = 181.475 uV 10:03:38 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 171.496 uVs 10:03:42 I1Y Average and RMS of Measurements VT = -182.887 +- 3.890 uVs I1Y = -36.575 +- 0.778 uTm 10:03:47 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:03:50 Integrated Voltage Measurement... 10:03:59 Voltage Offset Correction: v_off_ini = 153.728 uV, v_off_fin = 143.734 uV 10:03:59 Integrated Voltage Measurement dXYus = -1.0002 mm, dXYds = 0.0000 mm dXY = -1.0002 mm VT = -84.929 uVs 10:03:59 Integrated Voltage Measurement... 10:04:08 Voltage Offset Correction: v_off_ini = 145.428 uV, v_off_fin = 190.021 uV 10:04:08 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 99.147 uVs 10:04:08 Integrated Voltage Measurement... 10:04:17 Voltage Offset Correction: v_off_ini = 178.107 uV, v_off_fin = 162.041 uV 10:04:17 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -110.206 uVs 10:04:17 Integrated Voltage Measurement... 10:04:26 Voltage Offset Correction: v_off_ini = 161.931 uV, v_off_fin = 147.477 uV 10:04:26 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 123.107 uVs 10:04:26 Integrated Voltage Measurement... 10:04:35 Voltage Offset Correction: v_off_ini = 181.404 uV, v_off_fin = 171.961 uV 10:04:35 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -82.301 uVs 10:04:35 Integrated Voltage Measurement... 10:04:44 Voltage Offset Correction: v_off_ini = 168.141 uV, v_off_fin = 203.793 uV 10:04:44 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 94.074 uVs 10:04:46 I2X Average and RMS of Measurements VT = -98.961 +- 12.611 uVs I2X = 9.104 +- 1.161 uTm^2 10:04:51 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:04:54 Integrated Voltage Measurement... 10:05:03 Voltage Offset Correction: v_off_ini = 191.398 uV, v_off_fin = 179.770 uV 10:05:03 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -253.127 uVs 10:05:03 Integrated Voltage Measurement... 10:05:12 Voltage Offset Correction: v_off_ini = 186.921 uV, v_off_fin = 179.328 uV 10:05:12 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 249.812 uVs 10:05:12 Integrated Voltage Measurement... 10:05:21 Voltage Offset Correction: v_off_ini = 179.906 uV, v_off_fin = 153.084 uV 10:05:21 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -263.644 uVs 10:05:21 Integrated Voltage Measurement... 10:05:30 Voltage Offset Correction: v_off_ini = 164.932 uV, v_off_fin = 165.433 uV 10:05:30 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 233.245 uVs 10:05:30 Integrated Voltage Measurement... 10:05:39 Voltage Offset Correction: v_off_ini = 178.550 uV, v_off_fin = 166.171 uV 10:05:39 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -262.218 uVs 10:05:39 Integrated Voltage Measurement... 10:05:48 Voltage Offset Correction: v_off_ini = 162.472 uV, v_off_fin = 176.279 uV 10:05:48 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = 274.895 uVs 10:05:51 I2Y Average and RMS of Measurements VT = -256.157 +- 8.855 uVs I2Y = -23.567 +- 0.815 uTm^2 10:06:00 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:06:04 Integrated Voltage Measurement... 10:06:13 Voltage Offset Correction: v_off_ini = 217.587 uV, v_off_fin = 207.987 uV 10:06:14 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -146.560 uVs 10:06:14 Integrated Voltage Measurement... 10:06:22 Voltage Offset Correction: v_off_ini = 147.532 uV, v_off_fin = 165.917 uV 10:06:23 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 113.640 uVs 10:06:23 Integrated Voltage Measurement... 10:06:32 Voltage Offset Correction: v_off_ini = 190.235 uV, v_off_fin = 186.500 uV 10:06:32 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -77.774 uVs 10:06:32 Integrated Voltage Measurement... 10:06:41 Voltage Offset Correction: v_off_ini = 211.400 uV, v_off_fin = 197.257 uV 10:06:41 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 52.405 uVs 10:06:41 Integrated Voltage Measurement... 10:06:50 Voltage Offset Correction: v_off_ini = 183.684 uV, v_off_fin = 188.227 uV 10:06:50 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -76.659 uVs 10:06:50 Integrated Voltage Measurement... 10:06:59 Voltage Offset Correction: v_off_ini = 184.987 uV, v_off_fin = 183.287 uV 10:06:59 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 120.560 uVs 10:07:03 I1X Average and RMS of Measurements VT = -97.933 +- 26.545 uVs I1X = 19.587 +- 5.310 uTm 10:07:05 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:07:10 Integrated Voltage Measurement... 10:07:19 Voltage Offset Correction: v_off_ini = 202.849 uV, v_off_fin = 196.398 uV 10:07:19 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = -209.659 uVs 10:07:19 Integrated Voltage Measurement... 10:07:28 Voltage Offset Correction: v_off_ini = 217.383 uV, v_off_fin = 226.054 uV 10:07:28 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 174.350 uVs 10:07:28 Integrated Voltage Measurement... 10:07:37 Voltage Offset Correction: v_off_ini = 227.446 uV, v_off_fin = 204.407 uV 10:07:37 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -192.849 uVs 10:07:37 Integrated Voltage Measurement... 10:07:46 Voltage Offset Correction: v_off_ini = 236.402 uV, v_off_fin = 231.680 uV 10:07:46 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 183.848 uVs 10:07:46 Integrated Voltage Measurement... 10:07:55 Voltage Offset Correction: v_off_ini = 223.442 uV, v_off_fin = 218.805 uV 10:07:55 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -201.085 uVs 10:07:55 Integrated Voltage Measurement... 10:08:04 Voltage Offset Correction: v_off_ini = 236.394 uV, v_off_fin = 248.953 uV 10:08:04 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 210.109 uVs 10:08:08 I1Y Average and RMS of Measurements VT = -195.317 +- 7.421 uVs I1Y = -39.061 +- 1.483 uTm 10:08:13 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:08:16 Integrated Voltage Measurement... 10:08:25 Voltage Offset Correction: v_off_ini = 224.661 uV, v_off_fin = 238.325 uV 10:08:25 Integrated Voltage Measurement dXYus = -1.0002 mm, dXYds = 0.0000 mm dXY = -1.0002 mm VT = -97.698 uVs 10:08:25 Integrated Voltage Measurement... 10:08:34 Voltage Offset Correction: v_off_ini = 258.018 uV, v_off_fin = 252.722 uV 10:08:34 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 95.995 uVs 10:08:34 Integrated Voltage Measurement... 10:08:43 Voltage Offset Correction: v_off_ini = 262.312 uV, v_off_fin = 249.387 uV 10:08:43 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -123.548 uVs 10:08:43 Integrated Voltage Measurement... 10:08:52 Voltage Offset Correction: v_off_ini = 240.554 uV, v_off_fin = 246.746 uV 10:08:52 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 105.580 uVs 10:08:52 Integrated Voltage Measurement... 10:09:01 Voltage Offset Correction: v_off_ini = 243.730 uV, v_off_fin = 231.286 uV 10:09:01 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -85.063 uVs 10:09:01 Integrated Voltage Measurement... 10:09:10 Voltage Offset Correction: v_off_ini = 231.609 uV, v_off_fin = 250.844 uV 10:09:10 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 94.580 uVs 10:09:12 I2X Average and RMS of Measurements VT = -100.411 +- 10.411 uVs I2X = 9.238 +- 0.958 uTm^2 10:09:17 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:09:20 Integrated Voltage Measurement... 10:09:29 Voltage Offset Correction: v_off_ini = 235.703 uV, v_off_fin = 225.101 uV 10:09:29 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -236.884 uVs 10:09:29 Integrated Voltage Measurement... 10:09:38 Voltage Offset Correction: v_off_ini = 234.863 uV, v_off_fin = 226.857 uV 10:09:38 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 252.704 uVs 10:09:38 Integrated Voltage Measurement... 10:09:47 Voltage Offset Correction: v_off_ini = 243.779 uV, v_off_fin = 229.960 uV 10:09:47 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -272.277 uVs 10:09:47 Integrated Voltage Measurement... 10:09:56 Voltage Offset Correction: v_off_ini = 219.971 uV, v_off_fin = 228.356 uV 10:09:56 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 266.080 uVs 10:09:56 Integrated Voltage Measurement... 10:10:05 Voltage Offset Correction: v_off_ini = 220.117 uV, v_off_fin = 245.157 uV 10:10:05 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -260.480 uVs 10:10:05 Integrated Voltage Measurement... 10:10:14 Voltage Offset Correction: v_off_ini = 226.321 uV, v_off_fin = 234.189 uV 10:10:14 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 264.205 uVs 10:10:17 I2Y Average and RMS of Measurements VT = -258.772 +- 10.270 uVs I2Y = -23.807 +- 0.945 uTm^2 10:10:26 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:10:31 Integrated Voltage Measurement... 10:10:40 Voltage Offset Correction: v_off_ini = 219.115 uV, v_off_fin = 216.580 uV 10:10:40 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.4999 mm dXY = -0.4999 mm VT = -113.323 uVs 10:10:40 Integrated Voltage Measurement... 10:10:49 Voltage Offset Correction: v_off_ini = 213.685 uV, v_off_fin = 228.788 uV 10:10:49 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = 96.213 uVs 10:10:49 Integrated Voltage Measurement... 10:10:58 Voltage Offset Correction: v_off_ini = 233.119 uV, v_off_fin = 217.695 uV 10:10:58 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -81.934 uVs 10:10:58 Integrated Voltage Measurement... 10:11:07 Voltage Offset Correction: v_off_ini = 215.235 uV, v_off_fin = 226.984 uV 10:11:07 Integrated Voltage Measurement dXYus = 0.4999 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 85.527 uVs 10:11:07 Integrated Voltage Measurement... 10:11:16 Voltage Offset Correction: v_off_ini = 217.408 uV, v_off_fin = 244.268 uV 10:11:16 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -49.685 uVs 10:11:16 Integrated Voltage Measurement... 10:11:25 Voltage Offset Correction: v_off_ini = 263.416 uV, v_off_fin = 240.655 uV 10:11:25 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 90.366 uVs 10:11:29 I1X Average and RMS of Measurements VT = -86.175 +- 14.288 uVs I1X = 17.237 +- 2.859 uTm 10:11:31 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:11:36 Integrated Voltage Measurement... 10:11:45 Voltage Offset Correction: v_off_ini = 250.001 uV, v_off_fin = 232.380 uV 10:11:45 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = -203.022 uVs 10:11:45 Integrated Voltage Measurement... 10:11:54 Voltage Offset Correction: v_off_ini = 254.583 uV, v_off_fin = 237.364 uV 10:11:54 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 190.165 uVs 10:11:54 Integrated Voltage Measurement... 10:12:03 Voltage Offset Correction: v_off_ini = 226.386 uV, v_off_fin = 219.650 uV 10:12:03 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -181.377 uVs 10:12:03 Integrated Voltage Measurement... 10:12:12 Voltage Offset Correction: v_off_ini = 223.750 uV, v_off_fin = 238.602 uV 10:12:12 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 199.771 uVs 10:12:12 Integrated Voltage Measurement... 10:12:21 Voltage Offset Correction: v_off_ini = 216.260 uV, v_off_fin = 208.643 uV 10:12:21 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -184.033 uVs 10:12:21 Integrated Voltage Measurement... 10:12:30 Voltage Offset Correction: v_off_ini = 240.478 uV, v_off_fin = 245.780 uV 10:12:30 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 179.063 uVs 10:12:34 I1Y Average and RMS of Measurements VT = -189.572 +- 6.183 uVs I1Y = -37.913 +- 1.238 uTm 10:12:39 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:12:42 Integrated Voltage Measurement... 10:12:51 Voltage Offset Correction: v_off_ini = 193.667 uV, v_off_fin = 223.438 uV 10:12:51 Integrated Voltage Measurement dXYus = -1.0002 mm, dXYds = 0.0000 mm dXY = -1.0002 mm VT = -98.700 uVs 10:12:51 Integrated Voltage Measurement... 10:13:00 Voltage Offset Correction: v_off_ini = 238.899 uV, v_off_fin = 250.643 uV 10:13:00 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 92.557 uVs 10:13:00 Integrated Voltage Measurement... 10:13:09 Voltage Offset Correction: v_off_ini = 244.249 uV, v_off_fin = 227.867 uV 10:13:09 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -88.406 uVs 10:13:09 Integrated Voltage Measurement... 10:13:18 Voltage Offset Correction: v_off_ini = 251.219 uV, v_off_fin = 259.258 uV 10:13:18 Integrated Voltage Measurement dXYus = 0.9998 mm, dXYds = 0.0000 mm dXY = 0.9998 mm VT = 105.627 uVs 10:13:18 Integrated Voltage Measurement... 10:13:27 Voltage Offset Correction: v_off_ini = 241.630 uV, v_off_fin = 237.818 uV 10:13:27 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -93.714 uVs 10:13:27 Integrated Voltage Measurement... 10:13:36 Voltage Offset Correction: v_off_ini = 232.126 uV, v_off_fin = 250.118 uV 10:13:36 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 64.146 uVs 10:13:38 I2X Average and RMS of Measurements VT = -90.525 +- 8.219 uVs I2X = 8.328 +- 0.756 uTm^2 10:13:43 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:13:46 Integrated Voltage Measurement... 10:13:55 Voltage Offset Correction: v_off_ini = 232.361 uV, v_off_fin = 241.398 uV 10:13:55 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -257.894 uVs 10:13:55 Integrated Voltage Measurement... 10:14:04 Voltage Offset Correction: v_off_ini = 227.178 uV, v_off_fin = 247.816 uV 10:14:04 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 260.073 uVs 10:14:04 Integrated Voltage Measurement... 10:14:13 Voltage Offset Correction: v_off_ini = 254.387 uV, v_off_fin = 255.842 uV 10:14:13 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -265.457 uVs 10:14:13 Integrated Voltage Measurement... 10:14:22 Voltage Offset Correction: v_off_ini = 258.356 uV, v_off_fin = 253.693 uV 10:14:22 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 278.900 uVs 10:14:22 Integrated Voltage Measurement... 10:14:31 Voltage Offset Correction: v_off_ini = 237.216 uV, v_off_fin = 234.412 uV 10:14:31 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -279.321 uVs 10:14:31 Integrated Voltage Measurement... 10:14:40 Voltage Offset Correction: v_off_ini = 229.710 uV, v_off_fin = 233.221 uV 10:14:40 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 257.063 uVs 10:14:43 I2Y Average and RMS of Measurements VT = -266.451 +- 5.526 uVs I2Y = -24.514 +- 0.508 uTm^2 10:14:52 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:14:57 Integrated Voltage Measurement... 10:15:06 Voltage Offset Correction: v_off_ini = 216.204 uV, v_off_fin = 225.835 uV 10:15:06 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -88.964 uVs 10:15:06 Integrated Voltage Measurement... 10:15:15 Voltage Offset Correction: v_off_ini = 255.747 uV, v_off_fin = 232.573 uV 10:15:15 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 85.958 uVs 10:15:15 Integrated Voltage Measurement... 10:15:24 Voltage Offset Correction: v_off_ini = 206.048 uV, v_off_fin = 213.541 uV 10:15:24 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -103.588 uVs 10:15:24 Integrated Voltage Measurement... 10:15:33 Voltage Offset Correction: v_off_ini = 222.608 uV, v_off_fin = 237.512 uV 10:15:33 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 85.648 uVs 10:15:33 Integrated Voltage Measurement... 10:15:42 Voltage Offset Correction: v_off_ini = 231.387 uV, v_off_fin = 230.346 uV 10:15:42 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -88.026 uVs 10:15:42 Integrated Voltage Measurement... 10:15:51 Voltage Offset Correction: v_off_ini = 234.940 uV, v_off_fin = 209.014 uV 10:15:51 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 104.557 uVs 10:15:55 I1X Average and RMS of Measurements VT = -92.790 +- 3.830 uVs I1X = 18.558 +- 0.766 uTm 10:15:57 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:16:02 Integrated Voltage Measurement... 10:16:11 Voltage Offset Correction: v_off_ini = 212.365 uV, v_off_fin = 211.985 uV 10:16:11 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -186.628 uVs 10:16:11 Integrated Voltage Measurement... 10:16:20 Voltage Offset Correction: v_off_ini = 227.417 uV, v_off_fin = 233.383 uV 10:16:20 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 187.272 uVs 10:16:20 Integrated Voltage Measurement... 10:16:29 Voltage Offset Correction: v_off_ini = 229.678 uV, v_off_fin = 218.675 uV 10:16:29 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -192.850 uVs 10:16:29 Integrated Voltage Measurement... 10:16:38 Voltage Offset Correction: v_off_ini = 239.489 uV, v_off_fin = 251.614 uV 10:16:38 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 195.581 uVs 10:16:38 Integrated Voltage Measurement... 10:16:47 Voltage Offset Correction: v_off_ini = 241.981 uV, v_off_fin = 245.982 uV 10:16:47 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -203.630 uVs 10:16:47 Integrated Voltage Measurement... 10:16:56 Voltage Offset Correction: v_off_ini = 270.054 uV, v_off_fin = 272.403 uV 10:16:56 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 192.497 uVs 10:17:00 I1Y Average and RMS of Measurements VT = -193.076 +- 4.608 uVs I1Y = -38.615 +- 0.922 uTm 10:17:05 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:17:08 Integrated Voltage Measurement... 10:17:17 Voltage Offset Correction: v_off_ini = 260.654 uV, v_off_fin = 266.796 uV 10:17:17 Integrated Voltage Measurement dXYus = -1.0002 mm, dXYds = 0.0000 mm dXY = -1.0002 mm VT = -87.156 uVs 10:17:17 Integrated Voltage Measurement... 10:17:26 Voltage Offset Correction: v_off_ini = 270.065 uV, v_off_fin = 284.513 uV 10:17:26 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 70.572 uVs 10:17:26 Integrated Voltage Measurement... 10:17:35 Voltage Offset Correction: v_off_ini = 258.467 uV, v_off_fin = 265.700 uV 10:17:35 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -86.353 uVs 10:17:35 Integrated Voltage Measurement... 10:17:44 Voltage Offset Correction: v_off_ini = 274.064 uV, v_off_fin = 294.984 uV 10:17:44 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 104.649 uVs 10:17:44 Integrated Voltage Measurement... 10:17:53 Voltage Offset Correction: v_off_ini = 285.117 uV, v_off_fin = 279.939 uV 10:17:53 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -96.603 uVs 10:17:53 Integrated Voltage Measurement... 10:18:02 Voltage Offset Correction: v_off_ini = 296.631 uV, v_off_fin = 299.469 uV 10:18:02 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 103.347 uVs 10:18:04 I2X Average and RMS of Measurements VT = -91.447 +- 9.083 uVs I2X = 8.413 +- 0.836 uTm^2 10:18:09 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:18:12 Integrated Voltage Measurement... 10:18:21 Voltage Offset Correction: v_off_ini = 316.580 uV, v_off_fin = 325.511 uV 10:18:21 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -260.662 uVs 10:18:21 Integrated Voltage Measurement... 10:18:30 Voltage Offset Correction: v_off_ini = 337.261 uV, v_off_fin = 327.873 uV 10:18:30 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 260.288 uVs 10:18:30 Integrated Voltage Measurement... 10:18:39 Voltage Offset Correction: v_off_ini = 328.429 uV, v_off_fin = 320.878 uV 10:18:39 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -281.805 uVs 10:18:39 Integrated Voltage Measurement... 10:18:48 Voltage Offset Correction: v_off_ini = 322.526 uV, v_off_fin = 315.468 uV 10:18:48 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 279.877 uVs 10:18:48 Integrated Voltage Measurement... 10:18:57 Voltage Offset Correction: v_off_ini = 319.548 uV, v_off_fin = 322.995 uV 10:18:57 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -259.392 uVs 10:18:57 Integrated Voltage Measurement... 10:19:06 Voltage Offset Correction: v_off_ini = 319.662 uV, v_off_fin = 311.864 uV 10:19:06 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 242.019 uVs 10:19:09 I2Y Average and RMS of Measurements VT = -264.007 +- 12.554 uVs I2Y = -24.289 +- 1.155 uTm^2 10:19:18 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:19:22 Integrated Voltage Measurement... 10:19:31 Voltage Offset Correction: v_off_ini = 343.729 uV, v_off_fin = 325.141 uV 10:19:31 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -76.383 uVs 10:19:31 Integrated Voltage Measurement... 10:19:40 Voltage Offset Correction: v_off_ini = 327.026 uV, v_off_fin = 326.018 uV 10:19:40 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = 101.816 uVs 10:19:40 Integrated Voltage Measurement... 10:19:49 Voltage Offset Correction: v_off_ini = 328.271 uV, v_off_fin = 336.163 uV 10:19:49 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -75.717 uVs 10:19:49 Integrated Voltage Measurement... 10:19:58 Voltage Offset Correction: v_off_ini = 316.385 uV, v_off_fin = 309.289 uV 10:19:58 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = 97.593 uVs 10:19:58 Integrated Voltage Measurement... 10:20:07 Voltage Offset Correction: v_off_ini = 309.947 uV, v_off_fin = 304.174 uV 10:20:07 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -79.963 uVs 10:20:07 Integrated Voltage Measurement... 10:20:16 Voltage Offset Correction: v_off_ini = 318.958 uV, v_off_fin = 335.993 uV 10:20:16 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 95.580 uVs 10:20:21 I1X Average and RMS of Measurements VT = -87.842 +- 0.999 uVs I1X = 17.569 +- 0.199 uTm 10:20:23 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:20:27 Integrated Voltage Measurement... 10:20:36 Voltage Offset Correction: v_off_ini = 309.571 uV, v_off_fin = 328.545 uV 10:20:36 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = -189.081 uVs 10:20:36 Integrated Voltage Measurement... 10:20:45 Voltage Offset Correction: v_off_ini = 322.296 uV, v_off_fin = 349.253 uV 10:20:45 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 225.346 uVs 10:20:45 Integrated Voltage Measurement... 10:20:54 Voltage Offset Correction: v_off_ini = 344.348 uV, v_off_fin = 328.345 uV 10:20:54 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -195.639 uVs 10:20:54 Integrated Voltage Measurement... 10:21:03 Voltage Offset Correction: v_off_ini = 339.307 uV, v_off_fin = 348.147 uV 10:21:03 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 189.289 uVs 10:21:03 Integrated Voltage Measurement... 10:21:12 Voltage Offset Correction: v_off_ini = 343.989 uV, v_off_fin = 328.315 uV 10:21:12 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -189.630 uVs 10:21:12 Integrated Voltage Measurement... 10:21:21 Voltage Offset Correction: v_off_ini = 361.084 uV, v_off_fin = 357.010 uV 10:21:21 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 191.380 uVs 10:21:25 I1Y Average and RMS of Measurements VT = -196.727 +- 7.458 uVs I1Y = -39.346 +- 1.492 uTm 10:21:30 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:21:33 Integrated Voltage Measurement... 10:21:42 Voltage Offset Correction: v_off_ini = 371.852 uV, v_off_fin = 351.159 uV 10:21:42 Integrated Voltage Measurement dXYus = -1.0001 mm, dXYds = 0.0000 mm dXY = -1.0001 mm VT = -94.413 uVs 10:21:42 Integrated Voltage Measurement... 10:21:51 Voltage Offset Correction: v_off_ini = 347.583 uV, v_off_fin = 355.570 uV 10:21:51 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 90.558 uVs 10:21:51 Integrated Voltage Measurement... 10:22:00 Voltage Offset Correction: v_off_ini = 345.488 uV, v_off_fin = 343.609 uV 10:22:00 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -92.405 uVs 10:22:00 Integrated Voltage Measurement... 10:22:09 Voltage Offset Correction: v_off_ini = 352.658 uV, v_off_fin = 362.191 uV 10:22:09 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 81.424 uVs 10:22:09 Integrated Voltage Measurement... 10:22:18 Voltage Offset Correction: v_off_ini = 381.127 uV, v_off_fin = 368.843 uV 10:22:18 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -84.465 uVs 10:22:18 Integrated Voltage Measurement... 10:22:27 Voltage Offset Correction: v_off_ini = 365.302 uV, v_off_fin = 406.654 uV 10:22:27 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 73.323 uVs 10:22:29 I2X Average and RMS of Measurements VT = -86.098 +- 5.579 uVs I2X = 7.921 +- 0.513 uTm^2 10:22:34 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:22:37 Integrated Voltage Measurement... 10:22:46 Voltage Offset Correction: v_off_ini = 420.968 uV, v_off_fin = 409.059 uV 10:22:46 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -250.764 uVs 10:22:46 Integrated Voltage Measurement... 10:22:55 Voltage Offset Correction: v_off_ini = 396.621 uV, v_off_fin = 407.205 uV 10:22:55 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 257.772 uVs 10:22:55 Integrated Voltage Measurement... 10:23:04 Voltage Offset Correction: v_off_ini = 394.413 uV, v_off_fin = 389.717 uV 10:23:04 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -245.888 uVs 10:23:04 Integrated Voltage Measurement... 10:23:13 Voltage Offset Correction: v_off_ini = 383.428 uV, v_off_fin = 409.887 uV 10:23:13 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 246.582 uVs 10:23:13 Integrated Voltage Measurement... 10:23:22 Voltage Offset Correction: v_off_ini = 407.232 uV, v_off_fin = 415.123 uV 10:23:22 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -243.675 uVs 10:23:22 Integrated Voltage Measurement... 10:23:31 Voltage Offset Correction: v_off_ini = 406.559 uV, v_off_fin = 412.635 uV 10:23:31 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 267.320 uVs 10:23:34 I2Y Average and RMS of Measurements VT = -252.000 +- 4.107 uVs I2Y = -23.184 +- 0.378 uTm^2 10:23:43 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:23:47 Integrated Voltage Measurement... 10:23:56 Voltage Offset Correction: v_off_ini = 425.476 uV, v_off_fin = 413.585 uV 10:23:56 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -125.849 uVs 10:23:56 Integrated Voltage Measurement... 10:24:05 Voltage Offset Correction: v_off_ini = 429.163 uV, v_off_fin = 425.368 uV 10:24:05 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 112.782 uVs 10:24:05 Integrated Voltage Measurement... 10:24:14 Voltage Offset Correction: v_off_ini = 441.394 uV, v_off_fin = 418.049 uV 10:24:14 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -70.369 uVs 10:24:14 Integrated Voltage Measurement... 10:24:23 Voltage Offset Correction: v_off_ini = 433.559 uV, v_off_fin = 432.742 uV 10:24:23 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 65.625 uVs 10:24:23 Integrated Voltage Measurement... 10:24:32 Voltage Offset Correction: v_off_ini = 421.986 uV, v_off_fin = 400.641 uV 10:24:32 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -74.150 uVs 10:24:32 Integrated Voltage Measurement... 10:24:41 Voltage Offset Correction: v_off_ini = 401.831 uV, v_off_fin = 399.352 uV 10:24:41 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 86.969 uVs 10:24:46 I1X Average and RMS of Measurements VT = -89.290 +- 21.842 uVs I1X = 17.859 +- 4.369 uTm 10:24:48 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:24:52 Integrated Voltage Measurement... 10:25:01 Voltage Offset Correction: v_off_ini = 411.842 uV, v_off_fin = 413.997 uV 10:25:01 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = -193.595 uVs 10:25:01 Integrated Voltage Measurement... 10:25:10 Voltage Offset Correction: v_off_ini = 438.097 uV, v_off_fin = 446.324 uV 10:25:10 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 205.603 uVs 10:25:10 Integrated Voltage Measurement... 10:25:19 Voltage Offset Correction: v_off_ini = 398.342 uV, v_off_fin = 380.205 uV 10:25:19 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -194.764 uVs 10:25:19 Integrated Voltage Measurement... 10:25:28 Voltage Offset Correction: v_off_ini = 416.086 uV, v_off_fin = 428.385 uV 10:25:28 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 172.755 uVs 10:25:28 Integrated Voltage Measurement... 10:25:37 Voltage Offset Correction: v_off_ini = 421.755 uV, v_off_fin = 393.600 uV 10:25:37 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -179.435 uVs 10:25:37 Integrated Voltage Measurement... 10:25:46 Voltage Offset Correction: v_off_ini = 419.733 uV, v_off_fin = 442.659 uV 10:25:46 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 198.381 uVs 10:25:50 I1Y Average and RMS of Measurements VT = -190.755 +- 6.597 uVs I1Y = -38.151 +- 1.320 uTm 10:25:56 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:25:58 Integrated Voltage Measurement... 10:26:07 Voltage Offset Correction: v_off_ini = 413.665 uV, v_off_fin = 388.635 uV 10:26:07 Integrated Voltage Measurement dXYus = -1.0002 mm, dXYds = 0.0000 mm dXY = -1.0002 mm VT = -76.480 uVs 10:26:07 Integrated Voltage Measurement... 10:26:16 Voltage Offset Correction: v_off_ini = 408.749 uV, v_off_fin = 431.771 uV 10:26:16 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 104.848 uVs 10:26:16 Integrated Voltage Measurement... 10:26:25 Voltage Offset Correction: v_off_ini = 414.445 uV, v_off_fin = 411.225 uV 10:26:25 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -66.171 uVs 10:26:25 Integrated Voltage Measurement... 10:26:34 Voltage Offset Correction: v_off_ini = 423.954 uV, v_off_fin = 423.794 uV 10:26:34 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 55.079 uVs 10:26:34 Integrated Voltage Measurement... 10:26:43 Voltage Offset Correction: v_off_ini = 397.300 uV, v_off_fin = 409.252 uV 10:26:43 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -69.450 uVs 10:26:43 Integrated Voltage Measurement... 10:26:52 Voltage Offset Correction: v_off_ini = 439.637 uV, v_off_fin = 437.962 uV 10:26:52 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 83.704 uVs 10:26:55 I2X Average and RMS of Measurements VT = -75.955 +- 12.271 uVs I2X = 6.988 +- 1.129 uTm^2 10:27:00 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:27:02 Integrated Voltage Measurement... 10:27:11 Voltage Offset Correction: v_off_ini = 432.840 uV, v_off_fin = 408.336 uV 10:27:11 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -245.817 uVs 10:27:11 Integrated Voltage Measurement... 10:27:20 Voltage Offset Correction: v_off_ini = 411.836 uV, v_off_fin = 420.587 uV 10:27:20 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 273.121 uVs 10:27:20 Integrated Voltage Measurement... 10:27:29 Voltage Offset Correction: v_off_ini = 433.829 uV, v_off_fin = 432.794 uV 10:27:29 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -246.541 uVs 10:27:29 Integrated Voltage Measurement... 10:27:38 Voltage Offset Correction: v_off_ini = 441.591 uV, v_off_fin = 431.548 uV 10:27:38 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 257.922 uVs 10:27:38 Integrated Voltage Measurement... 10:27:47 Voltage Offset Correction: v_off_ini = 433.757 uV, v_off_fin = 439.503 uV 10:27:47 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -258.354 uVs 10:27:47 Integrated Voltage Measurement... 10:27:56 Voltage Offset Correction: v_off_ini = 443.528 uV, v_off_fin = 441.745 uV 10:27:56 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 255.377 uVs 10:27:59 I2Y Average and RMS of Measurements VT = -256.189 +- 2.993 uVs I2Y = -23.569 +- 0.275 uTm^2 10:28:10 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:28:14 Integrated Voltage Measurement... 10:28:23 Voltage Offset Correction: v_off_ini = 431.854 uV, v_off_fin = 419.452 uV 10:28:23 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -88.640 uVs 10:28:23 Integrated Voltage Measurement... 10:28:32 Voltage Offset Correction: v_off_ini = 420.494 uV, v_off_fin = 406.554 uV 10:28:32 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 110.492 uVs 10:28:32 Integrated Voltage Measurement... 10:28:41 Voltage Offset Correction: v_off_ini = 417.621 uV, v_off_fin = 428.379 uV 10:28:41 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -89.610 uVs 10:28:41 Integrated Voltage Measurement... 10:28:50 Voltage Offset Correction: v_off_ini = 419.136 uV, v_off_fin = 410.958 uV 10:28:50 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 90.416 uVs 10:28:50 Integrated Voltage Measurement... 10:28:59 Voltage Offset Correction: v_off_ini = 436.981 uV, v_off_fin = 436.583 uV 10:28:59 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -66.340 uVs 10:28:59 Integrated Voltage Measurement... 10:29:08 Voltage Offset Correction: v_off_ini = 445.246 uV, v_off_fin = 430.577 uV 10:29:08 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 66.814 uVs 10:29:12 I1X Average and RMS of Measurements VT = -85.385 +- 13.860 uVs I1X = 17.077 +- 2.772 uTm 10:29:14 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:29:19 Integrated Voltage Measurement... 10:29:28 Voltage Offset Correction: v_off_ini = 426.124 uV, v_off_fin = 433.965 uV 10:29:28 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -194.303 uVs 10:29:28 Integrated Voltage Measurement... 10:29:37 Voltage Offset Correction: v_off_ini = 444.320 uV, v_off_fin = 451.068 uV 10:29:37 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 194.813 uVs 10:29:37 Integrated Voltage Measurement... 10:29:46 Voltage Offset Correction: v_off_ini = 479.090 uV, v_off_fin = 461.952 uV 10:29:46 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -203.865 uVs 10:29:46 Integrated Voltage Measurement... 10:29:55 Voltage Offset Correction: v_off_ini = 429.115 uV, v_off_fin = 457.724 uV 10:29:55 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 191.097 uVs 10:29:55 Integrated Voltage Measurement... 10:30:04 Voltage Offset Correction: v_off_ini = 441.559 uV, v_off_fin = 422.037 uV 10:30:04 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -193.470 uVs 10:30:04 Integrated Voltage Measurement... 10:30:13 Voltage Offset Correction: v_off_ini = 439.427 uV, v_off_fin = 445.635 uV 10:30:13 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 210.756 uVs 10:30:17 I1Y Average and RMS of Measurements VT = -198.051 +- 3.110 uVs I1Y = -39.608 +- 0.622 uTm 10:30:22 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:30:24 Integrated Voltage Measurement... 10:30:33 Voltage Offset Correction: v_off_ini = 434.381 uV, v_off_fin = 437.584 uV 10:30:33 Integrated Voltage Measurement dXYus = -1.0002 mm, dXYds = 0.0000 mm dXY = -1.0002 mm VT = -88.133 uVs 10:30:33 Integrated Voltage Measurement... 10:30:42 Voltage Offset Correction: v_off_ini = 440.928 uV, v_off_fin = 446.749 uV 10:30:42 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 76.942 uVs 10:30:43 Integrated Voltage Measurement... 10:30:51 Voltage Offset Correction: v_off_ini = 461.837 uV, v_off_fin = 470.236 uV 10:30:52 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -62.050 uVs 10:30:52 Integrated Voltage Measurement... 10:31:00 Voltage Offset Correction: v_off_ini = 481.802 uV, v_off_fin = 493.369 uV 10:31:01 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 105.861 uVs 10:31:01 Integrated Voltage Measurement... 10:31:09 Voltage Offset Correction: v_off_ini = 476.788 uV, v_off_fin = 459.268 uV 10:31:10 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -72.324 uVs 10:31:10 Integrated Voltage Measurement... 10:31:19 Voltage Offset Correction: v_off_ini = 480.395 uV, v_off_fin = 483.318 uV 10:31:19 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 114.625 uVs 10:31:21 I2X Average and RMS of Measurements VT = -86.656 +- 4.856 uVs I2X = 7.972 +- 0.447 uTm^2 10:31:26 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:31:29 Integrated Voltage Measurement... 10:31:38 Voltage Offset Correction: v_off_ini = 456.214 uV, v_off_fin = 457.457 uV 10:31:38 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -229.004 uVs 10:31:38 Integrated Voltage Measurement... 10:31:47 Voltage Offset Correction: v_off_ini = 476.117 uV, v_off_fin = 477.288 uV 10:31:47 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 257.498 uVs 10:31:47 Integrated Voltage Measurement... 10:31:56 Voltage Offset Correction: v_off_ini = 479.031 uV, v_off_fin = 467.275 uV 10:31:56 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = -243.551 uVs 10:31:56 Integrated Voltage Measurement... 10:32:05 Voltage Offset Correction: v_off_ini = 490.276 uV, v_off_fin = 498.168 uV 10:32:05 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 279.405 uVs 10:32:05 Integrated Voltage Measurement... 10:32:14 Voltage Offset Correction: v_off_ini = 486.297 uV, v_off_fin = 479.251 uV 10:32:14 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -250.915 uVs 10:32:14 Integrated Voltage Measurement... 10:32:23 Voltage Offset Correction: v_off_ini = 474.655 uV, v_off_fin = 478.659 uV 10:32:23 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 249.745 uVs 10:32:25 I2Y Average and RMS of Measurements VT = -251.686 +- 7.503 uVs I2Y = -23.156 +- 0.691 uTm^2 10:32:36 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:32:40 Integrated Voltage Measurement... 10:32:49 Voltage Offset Correction: v_off_ini = 458.460 uV, v_off_fin = 476.122 uV 10:32:49 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4998 mm dXY = -0.4999 mm VT = -63.338 uVs 10:32:49 Integrated Voltage Measurement... 10:32:58 Voltage Offset Correction: v_off_ini = 476.804 uV, v_off_fin = 483.464 uV 10:32:58 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 68.538 uVs 10:32:58 Integrated Voltage Measurement... 10:33:07 Voltage Offset Correction: v_off_ini = 471.931 uV, v_off_fin = 464.642 uV 10:33:07 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -84.841 uVs 10:33:07 Integrated Voltage Measurement... 10:33:16 Voltage Offset Correction: v_off_ini = 473.391 uV, v_off_fin = 477.593 uV 10:33:16 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 93.103 uVs 10:33:16 Integrated Voltage Measurement... 10:33:25 Voltage Offset Correction: v_off_ini = 497.444 uV, v_off_fin = 475.709 uV 10:33:25 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -111.709 uVs 10:33:25 Integrated Voltage Measurement... 10:33:34 Voltage Offset Correction: v_off_ini = 462.195 uV, v_off_fin = 453.773 uV 10:33:34 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 93.961 uVs 10:33:39 I1X Average and RMS of Measurements VT = -85.915 +- 15.218 uVs I1X = 17.183 +- 3.043 uTm 10:33:41 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:33:45 Integrated Voltage Measurement... 10:33:54 Voltage Offset Correction: v_off_ini = 478.745 uV, v_off_fin = 491.818 uV 10:33:54 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -187.948 uVs 10:33:54 Integrated Voltage Measurement... 10:34:03 Voltage Offset Correction: v_off_ini = 468.845 uV, v_off_fin = 452.297 uV 10:34:03 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 215.459 uVs 10:34:03 Integrated Voltage Measurement... 10:34:12 Voltage Offset Correction: v_off_ini = 467.173 uV, v_off_fin = 455.196 uV 10:34:12 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -207.995 uVs 10:34:12 Integrated Voltage Measurement... 10:34:21 Voltage Offset Correction: v_off_ini = 455.988 uV, v_off_fin = 469.967 uV 10:34:21 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 208.330 uVs 10:34:21 Integrated Voltage Measurement... 10:34:30 Voltage Offset Correction: v_off_ini = 452.979 uV, v_off_fin = 434.012 uV 10:34:30 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -196.559 uVs 10:34:30 Integrated Voltage Measurement... 10:34:39 Voltage Offset Correction: v_off_ini = 471.872 uV, v_off_fin = 478.290 uV 10:34:39 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 192.304 uVs 10:34:44 I1Y Average and RMS of Measurements VT = -201.433 +- 5.609 uVs I1Y = -40.285 +- 1.121 uTm 10:34:49 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:34:51 Integrated Voltage Measurement... 10:35:00 Voltage Offset Correction: v_off_ini = 431.311 uV, v_off_fin = 444.929 uV 10:35:00 Integrated Voltage Measurement dXYus = -1.0001 mm, dXYds = 0.0000 mm dXY = -1.0001 mm VT = -73.409 uVs 10:35:00 Integrated Voltage Measurement... 10:35:09 Voltage Offset Correction: v_off_ini = 461.941 uV, v_off_fin = 490.569 uV 10:35:09 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 82.897 uVs 10:35:09 Integrated Voltage Measurement... 10:35:18 Voltage Offset Correction: v_off_ini = 476.686 uV, v_off_fin = 487.060 uV 10:35:18 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -89.034 uVs 10:35:18 Integrated Voltage Measurement... 10:35:27 Voltage Offset Correction: v_off_ini = 476.309 uV, v_off_fin = 492.714 uV 10:35:27 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 62.181 uVs 10:35:28 Integrated Voltage Measurement... 10:35:37 Voltage Offset Correction: v_off_ini = 467.088 uV, v_off_fin = 466.017 uV 10:35:37 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -72.260 uVs 10:35:37 Integrated Voltage Measurement... 10:35:46 Voltage Offset Correction: v_off_ini = 462.026 uV, v_off_fin = 498.080 uV 10:35:46 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 108.619 uVs 10:35:48 I2X Average and RMS of Measurements VT = -81.400 +- 6.476 uVs I2X = 7.489 +- 0.596 uTm^2 10:35:53 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:35:56 Integrated Voltage Measurement... 10:36:05 Voltage Offset Correction: v_off_ini = 494.028 uV, v_off_fin = 496.101 uV 10:36:05 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -254.999 uVs 10:36:05 Integrated Voltage Measurement... 10:36:14 Voltage Offset Correction: v_off_ini = 488.342 uV, v_off_fin = 495.582 uV 10:36:14 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 262.027 uVs 10:36:14 Integrated Voltage Measurement... 10:36:23 Voltage Offset Correction: v_off_ini = 514.491 uV, v_off_fin = 500.371 uV 10:36:23 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -247.391 uVs 10:36:23 Integrated Voltage Measurement... 10:36:32 Voltage Offset Correction: v_off_ini = 516.756 uV, v_off_fin = 524.890 uV 10:36:32 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 269.046 uVs 10:36:32 Integrated Voltage Measurement... 10:36:41 Voltage Offset Correction: v_off_ini = 530.098 uV, v_off_fin = 521.115 uV 10:36:41 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -239.703 uVs 10:36:41 Integrated Voltage Measurement... 10:36:50 Voltage Offset Correction: v_off_ini = 525.902 uV, v_off_fin = 529.170 uV 10:36:50 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 267.517 uVs 10:36:52 I2Y Average and RMS of Measurements VT = -256.781 +- 2.245 uVs I2Y = -23.624 +- 0.207 uTm^2 10:37:05 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:37:10 Integrated Voltage Measurement... 10:37:19 Voltage Offset Correction: v_off_ini = 526.795 uV, v_off_fin = 535.122 uV 10:37:19 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -54.175 uVs 10:37:19 Integrated Voltage Measurement... 10:37:28 Voltage Offset Correction: v_off_ini = 541.169 uV, v_off_fin = 553.734 uV 10:37:28 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 84.088 uVs 10:37:28 Integrated Voltage Measurement... 10:37:37 Voltage Offset Correction: v_off_ini = 570.178 uV, v_off_fin = 567.900 uV 10:37:37 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -60.158 uVs 10:37:37 Integrated Voltage Measurement... 10:37:46 Voltage Offset Correction: v_off_ini = 536.493 uV, v_off_fin = 513.950 uV 10:37:46 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 107.769 uVs 10:37:46 Integrated Voltage Measurement... 10:37:55 Voltage Offset Correction: v_off_ini = 535.822 uV, v_off_fin = 551.877 uV 10:37:55 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -78.527 uVs 10:37:55 Integrated Voltage Measurement... 10:38:04 Voltage Offset Correction: v_off_ini = 542.592 uV, v_off_fin = 546.866 uV 10:38:04 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 98.875 uVs 10:38:08 I1X Average and RMS of Measurements VT = -80.599 +- 8.336 uVs I1X = 16.120 +- 1.667 uTm 10:38:11 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:38:15 Integrated Voltage Measurement... 10:38:24 Voltage Offset Correction: v_off_ini = 552.195 uV, v_off_fin = 528.942 uV 10:38:24 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -194.670 uVs 10:38:24 Integrated Voltage Measurement... 10:38:33 Voltage Offset Correction: v_off_ini = 561.075 uV, v_off_fin = 576.144 uV 10:38:33 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 208.880 uVs 10:38:33 Integrated Voltage Measurement... 10:38:42 Voltage Offset Correction: v_off_ini = 551.740 uV, v_off_fin = 554.066 uV 10:38:42 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -230.216 uVs 10:38:42 Integrated Voltage Measurement... 10:38:51 Voltage Offset Correction: v_off_ini = 578.368 uV, v_off_fin = 598.936 uV 10:38:51 Integrated Voltage Measurement dXYus = -0.4999 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 181.278 uVs 10:38:51 Integrated Voltage Measurement... 10:39:00 Voltage Offset Correction: v_off_ini = 564.945 uV, v_off_fin = 589.045 uV 10:39:00 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -209.881 uVs 10:39:00 Integrated Voltage Measurement... 10:39:09 Voltage Offset Correction: v_off_ini = 580.369 uV, v_off_fin = 601.662 uV 10:39:09 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 220.887 uVs 10:39:13 I1Y Average and RMS of Measurements VT = -207.635 +- 5.714 uVs I1Y = -41.527 +- 1.142 uTm 10:39:19 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:39:21 Integrated Voltage Measurement... 10:39:30 Voltage Offset Correction: v_off_ini = 582.693 uV, v_off_fin = 588.848 uV 10:39:30 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -96.610 uVs 10:39:30 Integrated Voltage Measurement... 10:39:39 Voltage Offset Correction: v_off_ini = 578.549 uV, v_off_fin = 559.730 uV 10:39:39 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 51.542 uVs 10:39:39 Integrated Voltage Measurement... 10:39:48 Voltage Offset Correction: v_off_ini = 587.886 uV, v_off_fin = 583.839 uV 10:39:48 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -96.742 uVs 10:39:48 Integrated Voltage Measurement... 10:39:57 Voltage Offset Correction: v_off_ini = 584.170 uV, v_off_fin = 561.746 uV 10:39:57 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 65.995 uVs 10:39:57 Integrated Voltage Measurement... 10:40:06 Voltage Offset Correction: v_off_ini = 553.864 uV, v_off_fin = 571.615 uV 10:40:06 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = -91.302 uVs 10:40:06 Integrated Voltage Measurement... 10:40:15 Voltage Offset Correction: v_off_ini = 564.200 uV, v_off_fin = 538.961 uV 10:40:15 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = 115.952 uVs 10:40:18 I2X Average and RMS of Measurements VT = -86.357 +- 12.569 uVs I2X = 7.945 +- 1.157 uTm^2 10:40:23 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:40:26 Integrated Voltage Measurement... 10:40:35 Voltage Offset Correction: v_off_ini = 568.786 uV, v_off_fin = 569.391 uV 10:40:35 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -245.002 uVs 10:40:35 Integrated Voltage Measurement... 10:40:44 Voltage Offset Correction: v_off_ini = 552.055 uV, v_off_fin = 541.706 uV 10:40:44 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 271.749 uVs 10:40:44 Integrated Voltage Measurement... 10:40:53 Voltage Offset Correction: v_off_ini = 552.245 uV, v_off_fin = 565.066 uV 10:40:53 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -254.047 uVs 10:40:53 Integrated Voltage Measurement... 10:41:02 Voltage Offset Correction: v_off_ini = 552.418 uV, v_off_fin = 567.311 uV 10:41:02 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 262.618 uVs 10:41:02 Integrated Voltage Measurement... 10:41:11 Voltage Offset Correction: v_off_ini = 546.714 uV, v_off_fin = 539.038 uV 10:41:11 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -248.348 uVs 10:41:11 Integrated Voltage Measurement... 10:41:20 Voltage Offset Correction: v_off_ini = 547.173 uV, v_off_fin = 530.106 uV 10:41:20 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 248.773 uVs 10:41:22 I2Y Average and RMS of Measurements VT = -255.090 +- 4.617 uVs I2Y = -23.468 +- 0.425 uTm^2 10:41:35 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:41:40 Integrated Voltage Measurement... 10:41:49 Voltage Offset Correction: v_off_ini = 543.369 uV, v_off_fin = 536.937 uV 10:41:49 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -84.025 uVs 10:41:49 Integrated Voltage Measurement... 10:41:58 Voltage Offset Correction: v_off_ini = 532.315 uV, v_off_fin = 541.148 uV 10:41:58 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 96.107 uVs 10:41:58 Integrated Voltage Measurement... 10:42:07 Voltage Offset Correction: v_off_ini = 538.920 uV, v_off_fin = 535.694 uV 10:42:07 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -99.171 uVs 10:42:07 Integrated Voltage Measurement... 10:42:16 Voltage Offset Correction: v_off_ini = 527.307 uV, v_off_fin = 526.385 uV 10:42:16 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 85.787 uVs 10:42:16 Integrated Voltage Measurement... 10:42:25 Voltage Offset Correction: v_off_ini = 556.407 uV, v_off_fin = 563.232 uV 10:42:25 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -73.102 uVs 10:42:25 Integrated Voltage Measurement... 10:42:34 Voltage Offset Correction: v_off_ini = 544.954 uV, v_off_fin = 557.522 uV 10:42:34 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 87.705 uVs 10:42:38 I1X Average and RMS of Measurements VT = -87.650 +- 5.218 uVs I1X = 17.530 +- 1.044 uTm 10:42:40 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:42:45 Integrated Voltage Measurement... 10:42:54 Voltage Offset Correction: v_off_ini = 531.996 uV, v_off_fin = 533.280 uV 10:42:54 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = -200.423 uVs 10:42:54 Integrated Voltage Measurement... 10:43:03 Voltage Offset Correction: v_off_ini = 562.596 uV, v_off_fin = 572.172 uV 10:43:03 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 214.421 uVs 10:43:03 Integrated Voltage Measurement... 10:43:12 Voltage Offset Correction: v_off_ini = 559.733 uV, v_off_fin = 534.985 uV 10:43:12 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -202.572 uVs 10:43:12 Integrated Voltage Measurement... 10:43:21 Voltage Offset Correction: v_off_ini = 569.159 uV, v_off_fin = 589.947 uV 10:43:21 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 209.783 uVs 10:43:21 Integrated Voltage Measurement... 10:43:30 Voltage Offset Correction: v_off_ini = 576.724 uV, v_off_fin = 555.280 uV 10:43:30 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -214.269 uVs 10:43:30 Integrated Voltage Measurement... 10:43:39 Voltage Offset Correction: v_off_ini = 576.822 uV, v_off_fin = 575.226 uV 10:43:39 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 210.860 uVs 10:43:43 I1Y Average and RMS of Measurements VT = -208.721 +- 2.764 uVs I1Y = -41.741 +- 0.553 uTm 10:43:48 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:43:51 Integrated Voltage Measurement... 10:44:00 Voltage Offset Correction: v_off_ini = 555.265 uV, v_off_fin = 557.591 uV 10:44:00 Integrated Voltage Measurement dXYus = -1.0002 mm, dXYds = 0.0000 mm dXY = -1.0002 mm VT = -90.125 uVs 10:44:00 Integrated Voltage Measurement... 10:44:09 Voltage Offset Correction: v_off_ini = 557.183 uV, v_off_fin = 595.710 uV 10:44:09 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 104.144 uVs 10:44:09 Integrated Voltage Measurement... 10:44:18 Voltage Offset Correction: v_off_ini = 590.874 uV, v_off_fin = 584.525 uV 10:44:18 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -60.230 uVs 10:44:18 Integrated Voltage Measurement... 10:44:27 Voltage Offset Correction: v_off_ini = 612.370 uV, v_off_fin = 602.203 uV 10:44:27 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 76.701 uVs 10:44:27 Integrated Voltage Measurement... 10:44:36 Voltage Offset Correction: v_off_ini = 586.775 uV, v_off_fin = 570.211 uV 10:44:36 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -72.906 uVs 10:44:36 Integrated Voltage Measurement... 10:44:45 Voltage Offset Correction: v_off_ini = 580.539 uV, v_off_fin = 573.009 uV 10:44:45 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 69.287 uVs 10:44:47 I2X Average and RMS of Measurements VT = -78.899 +- 12.939 uVs I2X = 7.258 +- 1.190 uTm^2 10:44:52 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:44:55 Integrated Voltage Measurement... 10:45:04 Voltage Offset Correction: v_off_ini = 595.355 uV, v_off_fin = 588.507 uV 10:45:04 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -256.786 uVs 10:45:04 Integrated Voltage Measurement... 10:45:13 Voltage Offset Correction: v_off_ini = 578.020 uV, v_off_fin = 591.317 uV 10:45:13 Integrated Voltage Measurement dXYus = -0.9999 mm, dXYds = 0.0000 mm dXY = -0.9999 mm VT = 238.527 uVs 10:45:13 Integrated Voltage Measurement... 10:45:22 Voltage Offset Correction: v_off_ini = 593.616 uV, v_off_fin = 599.913 uV 10:45:22 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -254.995 uVs 10:45:22 Integrated Voltage Measurement... 10:45:31 Voltage Offset Correction: v_off_ini = 609.714 uV, v_off_fin = 615.083 uV 10:45:31 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 244.145 uVs 10:45:31 Integrated Voltage Measurement... 10:45:40 Voltage Offset Correction: v_off_ini = 603.159 uV, v_off_fin = 610.640 uV 10:45:40 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -241.748 uVs 10:45:40 Integrated Voltage Measurement... 10:45:49 Voltage Offset Correction: v_off_ini = 610.311 uV, v_off_fin = 626.181 uV 10:45:49 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 268.146 uVs 10:45:52 I2Y Average and RMS of Measurements VT = -250.725 +- 3.086 uVs I2Y = -23.067 +- 0.284 uTm^2 10:46:05 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:46:09 Integrated Voltage Measurement... 10:46:18 Voltage Offset Correction: v_off_ini = 612.079 uV, v_off_fin = 601.483 uV 10:46:18 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.4999 mm dXY = -0.5000 mm VT = -69.041 uVs 10:46:18 Integrated Voltage Measurement... 10:46:27 Voltage Offset Correction: v_off_ini = 629.749 uV, v_off_fin = 629.944 uV 10:46:27 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 56.687 uVs 10:46:27 Integrated Voltage Measurement... 10:46:36 Voltage Offset Correction: v_off_ini = 626.856 uV, v_off_fin = 635.331 uV 10:46:36 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -72.403 uVs 10:46:36 Integrated Voltage Measurement... 10:46:45 Voltage Offset Correction: v_off_ini = 632.842 uV, v_off_fin = 637.599 uV 10:46:45 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 72.990 uVs 10:46:45 Integrated Voltage Measurement... 10:46:54 Voltage Offset Correction: v_off_ini = 634.743 uV, v_off_fin = 629.029 uV 10:46:54 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = -85.765 uVs 10:46:54 Integrated Voltage Measurement... 10:47:03 Voltage Offset Correction: v_off_ini = 652.747 uV, v_off_fin = 661.304 uV 10:47:03 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5000 mm dXY = 0.5000 mm VT = 57.888 uVs 10:47:07 I1X Average and RMS of Measurements VT = -69.129 +- 4.444 uVs I1X = 13.826 +- 0.888 uTm 10:47:10 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:47:14 Integrated Voltage Measurement... 10:47:23 Voltage Offset Correction: v_off_ini = 648.528 uV, v_off_fin = 646.942 uV 10:47:23 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.4999 mm dXY = 0.5000 mm VT = -216.094 uVs 10:47:23 Integrated Voltage Measurement... 10:47:32 Voltage Offset Correction: v_off_ini = 648.033 uV, v_off_fin = 640.327 uV 10:47:32 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 216.775 uVs 10:47:32 Integrated Voltage Measurement... 10:47:41 Voltage Offset Correction: v_off_ini = 645.828 uV, v_off_fin = 631.457 uV 10:47:41 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -222.573 uVs 10:47:41 Integrated Voltage Measurement... 10:47:50 Voltage Offset Correction: v_off_ini = 633.047 uV, v_off_fin = 635.141 uV 10:47:50 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5001 mm dXY = -0.5000 mm VT = 197.524 uVs 10:47:50 Integrated Voltage Measurement... 10:47:59 Voltage Offset Correction: v_off_ini = 611.096 uV, v_off_fin = 624.465 uV 10:47:59 Integrated Voltage Measurement dXYus = 0.5000 mm, dXYds = 0.5001 mm dXY = 0.5000 mm VT = -211.046 uVs 10:47:59 Integrated Voltage Measurement... 10:48:08 Voltage Offset Correction: v_off_ini = 628.663 uV, v_off_fin = 632.278 uV 10:48:08 Integrated Voltage Measurement dXYus = -0.5000 mm, dXYds = -0.5000 mm dXY = -0.5000 mm VT = 196.902 uVs 10:48:13 I1Y Average and RMS of Measurements VT = -210.152 +- 5.088 uVs I1Y = -42.028 +- 1.018 uTm 10:48:18 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:48:20 Integrated Voltage Measurement... 10:48:29 Voltage Offset Correction: v_off_ini = 598.497 uV, v_off_fin = 611.620 uV 10:48:29 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -74.613 uVs 10:48:29 Integrated Voltage Measurement... 10:48:38 Voltage Offset Correction: v_off_ini = 604.058 uV, v_off_fin = 634.100 uV 10:48:38 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 119.086 uVs 10:48:38 Integrated Voltage Measurement... 10:48:47 Voltage Offset Correction: v_off_ini = 619.984 uV, v_off_fin = 621.684 uV 10:48:47 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -96.886 uVs 10:48:47 Integrated Voltage Measurement... 10:48:56 Voltage Offset Correction: v_off_ini = 622.524 uV, v_off_fin = 640.295 uV 10:48:56 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 108.618 uVs 10:48:57 Integrated Voltage Measurement... 10:49:06 Voltage Offset Correction: v_off_ini = 626.153 uV, v_off_fin = 619.983 uV 10:49:06 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = -94.669 uVs 10:49:06 Integrated Voltage Measurement... 10:49:15 Voltage Offset Correction: v_off_ini = 624.379 uV, v_off_fin = 630.664 uV 10:49:15 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = 117.404 uVs 10:49:17 I2X Average and RMS of Measurements VT = -101.879 +- 3.801 uVs I2X = 9.373 +- 0.350 uTm^2 10:49:22 Wire Position: X0 = 0.0000 mm, Y0 = 0.0000 mm 10:49:25 Integrated Voltage Measurement... 10:49:34 Voltage Offset Correction: v_off_ini = 585.916 uV, v_off_fin = 602.355 uV 10:49:34 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0001 mm dXY = 1.0000 mm VT = -252.544 uVs 10:49:34 Integrated Voltage Measurement... 10:49:43 Voltage Offset Correction: v_off_ini = 576.036 uV, v_off_fin = 601.993 uV 10:49:43 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = -0.0001 mm dXY = -1.0000 mm VT = 275.243 uVs 10:49:43 Integrated Voltage Measurement... 10:49:52 Voltage Offset Correction: v_off_ini = 603.529 uV, v_off_fin = 584.723 uV 10:49:52 Integrated Voltage Measurement dXYus = 1.0000 mm, dXYds = 0.0000 mm dXY = 1.0000 mm VT = -281.456 uVs 10:49:52 Integrated Voltage Measurement... 10:50:01 Voltage Offset Correction: v_off_ini = 604.259 uV, v_off_fin = 596.926 uV 10:50:01 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 259.018 uVs 10:50:01 Integrated Voltage Measurement... 10:50:10 Voltage Offset Correction: v_off_ini = 602.763 uV, v_off_fin = 597.205 uV 10:50:10 Integrated Voltage Measurement dXYus = 0.9999 mm, dXYds = 0.0000 mm dXY = 0.9999 mm VT = -242.608 uVs 10:50:10 Integrated Voltage Measurement... 10:50:19 Voltage Offset Correction: v_off_ini = 613.729 uV, v_off_fin = 583.559 uV 10:50:19 Integrated Voltage Measurement dXYus = -1.0000 mm, dXYds = 0.0000 mm dXY = -1.0000 mm VT = 216.449 uVs 10:50:21 I2Y Average and RMS of Measurements VT = -254.553 +- 17.884 uVs I2Y = -23.419 +- 1.645 uTm^2